Introduction to
Pharmaceutical
Chemical Analysis
Introduction to
Pharmaceutical
Chemical Analysis
STEEN HANSEN
STIG PEDERSEN-BJERGAARD
KNUT RASMUSSEN
This edition first published 2012
Ó 2012 John Wiley & Sons Ltd.
Registered office
John Wiley & Sons Ltd, The Atrium, Southern Gate, Chichester, West Sussex, PO19 8SQ, United Kingdom
For details of our global editorial offices, for customer services and for information about how to apply for permission to reuse
the copyright material in this book please see our website at www.wiley.com.
The right of the author to be identified as the author of this work has been asserted in accordance with the Copyright,
Designs and Patents Act 1988.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form
or by any means, electronic, mechanical, photocopying, recording or otherwise, except as permitted by the UK Copyright,
Designs and Patents Act 1988, without the prior permission of the publisher.
Wiley also publishes its books in a variety of electronic formats. Some content that appears in print may not be available
in electronic books.
Designations used by companies to distinguish their products are often claimed as trademarks. All brand names and product
names used in this book are trade names, service marks, trademarks or registered trademarks of their respective owners.
The publisher is not associated with any product or vendor mentioned in this book. This publication is designed to provide
accurate and authoritative information in regard to the subject matter covered. It is sold on the understanding that the
publisher is not engaged in rendering professional services. If professional advice or other expert assistance is required,
the services of a competent professional should be sought.
The publisher and the author make no representations or warranties with respect to the accuracy or completeness of the
contents of this work and specifically disclaim all warranties, including without limitation any implied warranties of fitness for a
particular purpose. This work is sold with the understanding that the publisher is not engaged in rendering professional services.
The advice and strategies contained herein may not be suitable for every situation. In view of ongoing research, equipment
modifications, changes in governmental regulations, and the constant flow of information relating to the use of experimental
reagents, equipment, and devices, the reader is urged to review and evaluate the information provided in the package
insert or instructions for each chemical, piece of equipment, reagent, or device for, among other things, any changes in the
instructions or indication of usage and for added warnings and precautions. The fact that an organization or Website is referred
to in this work as a citation and/or a potential source of further information does not mean that the author or the publisher
endorses the information the organization or Website may provide or recommendations it may make. Further, readers should be
aware that Internet Websites listed in this work may have changed or disappeared between when this work was written and
when it is read. No warranty may be created or extended by any promotional statements for this work. Neither the publisher nor the
author shall be liable for any damages arising herefrom.
Library of Congress Cataloging-in-Publication Data
Hansen, Steen, 1947–
Chemical analysis in pharmaceutical sciences / Steen Hansen, Stig
Pedersen-Bjergaard, Knut Rasmussen.
p. ; cm.
Includes bibliographical references and index.
ISBN 978-0-470-66121-5 (cloth) – ISBN 978-0-470-66122-2 (pbk.)
1. Drugs–Analysis. 2. Pharmaceutical chemistry. I. Pedersen-Bjergaard,
Stig. II. Rasmussen, Knut. III. Title.
[DNLM: 1. Pharmaceutical Preparations–analysis. 2. Chemistry,
Pharmaceutical. 3. Drug Compounding–standards. QV 55]
RS189.H277 2012
615.1 0 9–dc23
2011030362
A catalogue record for this book is available from the British Library.
Print ISBN (Hardback): 9780470661215
Print ISBN (Paperback): 9780470661222
ePDF ISBN: 9781119953609
oBook ISBN: 9781119953647
ePub ISBN: 9781119954330
Mobi ISBN: 9781119954347
Set in 10/12pt Times Roman by Thomson Digital, Noida, India
Table of Contents
Preface
xv
1 Introduction to Pharmaceutical Analysis
1.1 Applications and Definitions
1.2 The Life of Medicines
1.3 The Quality of Medical Products
1.4 Summary
1
1
4
8
11
2 International Pharmacopoeias, Regulations and Guidelines
2.1 Overview of Legislation
2.2 Legislation and Regulations for Industrial Production
2.3 Life Time of Drugs and Drug Substances
2.4 Pharmacopoeias
2.5 International Harmonization
2.5.1 International Conference on Harmonization
2.5.2 Pharmacopoeial Discussion Group
2.6 Legislation and Regulations for Pharmacy Production
2.7 Summary
13
13
14
17
18
19
20
20
20
21
3 Fundamental Chemical Properties, Buffers and pH
3.1 pH and pKa
3.2 Partition
3.3 Stereochemistry
3.4 Stability Testing
3.5 Summary
23
23
25
28
29
30
4 Fundamentals of Pharmaceutical Analysis
4.1 What is a Pharmaceutical (Chemical) Analysis?
4.2 How to Specify Quantities and Concentrations?
4.3 Basic Laboratory Equipment
4.3.1 The Analytical Balance
4.3.2 Pipettes
4.3.3 Volumetric Flasks
4.3.4 Burettes
33
33
35
37
37
41
44
47
vi
Table of Contents
4.4 How to Make Solutions and Dilutions
4.5 Calibration of Analytical Methods
4.6 Errors, Accuracy, and Precision
4.6.1 Systematic and Random Errors
4.6.2 Accuracy and Precision
4.7 Statistics
4.7.1 Mean Value and Standard Deviation
4.7.2 Confidence Intervals
4.7.3 Comparison of Means with a t-Test
4.7.4 Q-Test to Reject Outliers
4.7.5 Linear Regression with the Method of Least Squares
4.7.6 How to Present an Analytical Result
4.8 Some Words and Concepts
4.8.1 Analysis and Determination
4.8.2 Sample Replicates and Measuring Replicates
4.8.3 Interference
4.8.4 Blind Samples
47
49
50
50
51
52
52
54
55
56
57
58
62
62
62
62
62
5 Titrimetric Methods
5.1 Introduction
5.2 Acid–Base Titrations
5.3 Acid–Base Titrations in Non-Aqueous Media
5.4 Redox Titrations
5.5 Other Principles of Titration
5.6 Summary
65
65
72
75
78
81
82
6 Introduction to Spectroscopic Methods
6.1 Electromagnetic Radiation
6.2 Molecules and Electromagnetic Radiation
6.3 Atoms and Electromagnetic Radiation
6.4 Summary
83
83
85
86
88
7 UV Spectrophotometry
7.1 Principle of Quantitative Determination
7.2 Principle of Identification
7.3 Which Substances Have Strong UV Absorbance?
7.4 Instrumentation
7.5 Practical Work and Method Development
7.6 Areas of Usage and Performance
7.7 System Testing
7.8 Summary
89
89
94
95
95
99
101
101
102
8 IR Spectrophotometry
8.1 IR Spectrophotometry
8.2 Instrumentation
8.3 Scope
103
103
106
109
Table of Contents
8.4 Instrument Calibration
8.5 NIR Spectrophotometry
8.6 Applications
8.7 Summary
vii
109
110
112
114
9 Atomic Spectrometry
9.1 Atomic Absorption Spectrometry
9.2 Instrumentation
9.3 Applications and Performance
9.4 Practical Work and Method Development
9.5 Atomic Emission Spectrometry
9.6 Instrumentation
9.7 Summary
115
115
118
121
122
123
124
124
10 Chromatography
10.1 General Principles
10.2 Retention
10.3 Column Efficiency
10.4 Selectivity
10.5 Peak Symmetry
10.6 Resolution
10.7 Chromatographic Techniques
10.8 Summary
127
127
131
133
135
136
138
140
140
11 Chromatographic Separation Principles
11.1 General Introduction
11.2 Normal Phase Chromatography
11.2.1 Silica
11.2.2 Interactions
11.2.3 Order of Elution
11.2.4 Other Stationary Phases
11.2.5 Mobile Phases
11.2.6 Summary of Normal Phase Chromatography
11.3 Reversed Phase Chromatography
11.3.1 Stationary Phases
11.3.2 Retention Mechanisms
11.3.3 Mobile Phases
11.3.4 Ion-Pair Chromatography
11.3.5 Summary of Reversed Phase Chromatography
11.4 Hydrophilic Interaction Chromatography
11.5 Chiral Separations
11.6 Size Exclusion Chromatography
11.6.1 Principle
11.6.2 Summary of SEC
11.7 Ion Exchange Chromatography
141
141
142
142
143
144
145
146
147
148
148
150
152
155
155
156
156
158
158
160
160
viii
Table of Contents
12 Thin-Layer Chromatography
12.1 Introduction
12.2 Apparatus
12.3 TLC Plates
12.4 Stationary Phases
12.5 Mobile Phases
12.6 Chromatographic Development
12.7 Detection
12.8 Applications of TLC
12.9 Quantitative Analysis and Instrumentation
12.10 Summary
163
163
164
166
166
167
168
169
169
170
171
13 High Performance Liquid Chromatography
13.1 Introduction
13.2 The Chromatographic Separation Process
13.3 The Column
13.4 Pumps
13.5 Detectors
13.5.1 UV detector
13.5.2 Fluorescence Detector
13.5.3 Electrochemical Detector
13.5.4 Refractive Index, Evaporative Light Scattering
and Corona Discharge Detectors
13.5.5 Combination of Detectors
13.6 Injectors
13.7 Mobile Phases
13.8 Solvents for Sample Preparation
13.9 Reporting the Results
13.10 Summary
173
173
175
177
180
182
182
184
186
14 Gas Chromatography
14.1 Introduction
14.2 Apparatus
14.3 Temperature
14.4 Carrier Gas
14.5 Stationary Phases
14.6 Selectivity in GC
14.7 Columns
14.7.1 Capillary Columns
14.7.2 Packed Columns
14.8 Injection Systems
14.8.1 Injection Systems for Capillary Columns
14.8.2 Injection Systems for Packed Columns
14.9 Detectors
14.9.1 Flame Ionization Detector
14.9.2 Nitrogen–Phosphorus Detector
191
191
192
193
195
196
197
198
198
199
200
200
202
203
203
203
186
187
187
188
189
189
190
Table of Contents
ix
14.9.3 Thermal Conductivity Detector
14.9.4 Electron Capture Detector
14.9.5 Mass Spectrometry Detector
14.10 Derivatization
14.10.1 Silylation
14.10.2 Alkylation
14.10.3 Acylation
14.11 The Uses of GC
14.12 More Advanced GC techniques
14.13 Summary
204
204
206
206
206
207
207
208
209
209
15 Capillary Electrophoresis
15.1 Principle and Theory
15.2 Electroosmotic Flow
15.3 Instrumentation
15.4 The Capillary
15.5 Sample Introduction
15.6 Capillary Zone Electrophoresis; an Example
15.7 Micellar Electrokinetic Chromatography
15.8 Chiral Separations
15.9 Coated Capillaries
15.10 Non-Aqueous CE
15.11 Summary
211
211
213
214
217
218
221
222
224
225
229
229
16 Mass Spectrometry
16.1 Introduction
16.2 Basic Theory
16.3 Electron Ionization
16.4 Identification using Electron Ionization Spectra
16.5 Characterization of Totally Unknowns using Electron
Ionization Spectra
16.6 Chemical Ionization
16.7 Electrospray Ionization
16.8 Atmospheric Pressure Chemical Ionization
16.9 High-Resolution Mass Spectrometry
16.10 Instrumentation
16.11 Chromatography Coupled with Mass Spectrometry
16.12 Quantitative GC-MS and LC-MS
16.13 Areas of Usage and Performance
16.14 Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry
16.15 Inductively Coupled Plasma Mass Spectrometry
16.16 Summary
231
231
233
236
237
239
244
246
247
248
250
253
256
257
257
258
259
17 Miscellaneous Chemical Techniques
17.1 Potentiometric Determination of Ions using Ion-Selective Electrodes
17.2 Paper Chromatography
261
261
263
x
Table of Contents
17.3 Supercritical Fluid Chromatography
17.4 Gel Electrophoresis
17.5 Iso-Electric Focusing
17.6 Nuclear Magnetic Resonance Spectrometry
17.7 Raman Spectrometry
264
265
267
268
270
18 Sample Preparation
18.1 Why is Sample Preparation Required?
18.2 Main Strategies
18.3 Recovery and Enrichment
18.4 Protein Precipitation
18.5 Liquid–Liquid Extraction
18.5.1 Fundamentals
18.5.2 A Closer Look at the Theory
18.5.3 Extraction Solvents
18.5.4 Calculation of Recovery
18.5.5 Multiple Extractions
18.5.6 LLE with Back-Extraction
18.6 Solid–Liquid Extraction
18.7 Solid Phase Extraction
18.7.1 Fundamentals
18.7.2 The SPE Column
18.7.3 Conditioning
18.7.4 Equipment
18.7.5 Reversed-Phase SPE
18.7.6 Secondary Interactions
18.7.7 Ion Exchange SPE
18.7.8 Mixed-Mode SPE
18.7.9 Normal-Phase SPE
18.8 Summary
273
273
274
276
278
279
279
279
282
283
285
286
287
287
287
288
289
290
290
292
293
295
297
298
19 Analytical Chemical Characteristics of Selected Drug Substances
19.1 Amitriptyline and Mianserin
19.2 Morphine and Codeine
19.3 Ibuprofen and Naproxen
19.4 Furosemide
19.5 Paracetamol (Acetaminophen)
19.6 Neutral Drugs
299
299
301
302
304
306
307
20 Quantification and Quality of Analytical Data
20.1 Peak Height and Peak Area
20.2 Calibration Methods
20.2.1 External Standard Method
20.2.2 Internal Standard Method
20.2.3 Standard Addition
20.2.4 Normalization
309
309
310
310
313
314
314
Table of Contents
20.3 Validation
20.3.1 Analytical Procedure
20.3.2 Accuracy
20.3.3 Precision
20.3.4 Specificity
20.3.5 Detection Limit
20.3.6 Quantification Limit
20.3.7 Linearity and Range
20.3.8 Robustness
20.3.9 Test Methods in the European Pharmacopeia
20.4 System Suitability
20.4.1 Adjustment of Chromatographic Conditions
21 Chemical Analysis of Drug Substances
21.1 What is a Pharmaceutical Raw Material, how is it Produced
and why must it be Controlled?
21.2 The Pharmacopoeias – the Basis for Control of Pharmaceutical
Raw Materials
21.3 Which Contaminants are Found in Raw Materials,
What are the Requirements in a Maximum Content and Why?
21.3.1 Well Defined Chemical Compounds
21.3.2 Mixtures of Organic Compounds
21.4 How to Check the Identity of Pharmaceutical Raw Materials
21.4.1 Overview of the Identification Procedures
21.4.2 Techniques used for the Identification of Well Defined
Chemical Compounds
21.4.2.1 Infrared Absorption Spectrophotometry
21.4.2.2 Ultraviolet and Visible Absorption
Spectrophotometry
21.4.2.3 Thin-Layer Chromatography
21.4.2.4 Melting Point
21.4.2.5 Polarimetry
21.4.2.6 High Performance Liquid Chromatography
21.4.2.7 Chloride and Sulfate Identification
21.5 How to Test for Impurities in Pharmaceutical Raw Materials
21.5.1 Main Purity Tests for Well Defined Chemical Compounds
21.5.1.1 Appearance of Solution
21.5.1.2 Absorbance
21.5.1.3 Acidity/Alkalinity
21.5.1.4 Optical Rotation
21.5.1.5 Related Substances
21.5.1.6 Solvent Residues
21.5.1.7 Foreign Anions
21.5.1.8 Cationic Impurities
21.5.1.9 Loss on Drying
21.5.1.10 Determination of Water
xi
314
317
317
318
320
320
321
321
323
325
325
326
327
327
330
337
339
343
344
344
344
344
347
351
352
353
356
359
359
359
361
364
365
365
366
372
372
376
378
379
xii
Table of Contents
21.5.2 Purity Tests for Raw Materials of the Type
of Mixtures of Organic Compounds
21.5.2.1 Oxidizing Substances
21.5.2.2 Acid Value
21.5.2.3 Hydroxyl Value
21.5.2.4 Iodine Value
21.5.2.5 Peroxide Value
21.5.2.6 Saponification Value
21.5.2.7 Unsaponifiable Matter
21.5.2.8 Other Tests
21.5.3 Identification of the Raw Materials of the Type of Mixtures
of Organic Compounds
21.6 How to Determine the Purity of Pharmaceutical Raw Materials
21.6.1 Acid–Base Titration in Aqueous Environment
21.6.2 Acid–Base Titration in a Non-Aqueous Environment
21.6.3 Redox Titrations
21.6.4 High Performance Liquid Chromatography
21.6.5 UV spectrophotometry
21.7 How to Control Compounds for Which no Pharmacopoeia
Monograph Exists
21.8 How are Ph.Eur. and USP Updated?
382
383
383
384
384
385
385
386
386
388
389
389
393
396
396
401
402
402
22 Chemical Analysis of Final Pharmaceutical Products
22.1 Quality Control of Final Pharmaceutical Products
22.2 Monographs and Chemical Testing
22.3 Identification of the Active Pharmaceutical Ingredient
22.4 Assay of the Active Pharmaceutical Ingredient
22.5 Chemical Tests for Final Pharmaceutical Products
22.5.1 Test for Related Substances
22.5.2 Uniformity of Content
22.5.3 Dissolution
405
405
406
412
427
446
446
449
451
23 Analysis of Drugs in Biological Fluids
23.1 Introduction
23.1.1 Drug Development
23.1.2 Therapeutic Drug Monitoring
23.1.3 Forensic and Toxicological Analysis
23.1.4 Doping Control Analysis
23.2 The Biological Matrix
23.3 Bioanalytical Methods
23.3.1 Sampling
23.3.2 Sample Preparation
23.3.3 Protein Precipitation
23.3.4 Liquid–Liquid Extraction
23.3.5 Solid-Phase Extraction
23.3.6 Separation
453
453
453
455
456
457
458
460
460
461
462
463
463
464
Table of Contents
23.3.7 Detection
23.3.8 Calibration and Quantification
23.4 Examples
23.4.1 Sample Preparation
23.4.1.1 Sample Preparation Procedure by LLE
23.4.1.2 Comments to the Procedure
23.4.1.3 Sample Preparation Procedure by LLE
and Back Extraction
23.4.1.4 Comments to the Procedure
23.4.1.5 Sample Preparation Procedure by SPE
23.4.1.6 Comments to the Procedure
23.4.1.7 Sample Preparation Procedure by Protein
Precipitation
23.4.1.8 Comments to the Procedure
23.4.2 Quantitative Determination
23.4.2.1 Quantitative Determination of Amitriptyline
in Serum by LC-MS
23.4.2.2 Comments to the Procedure
23.4.2.3 Determination of Valproic Acid in Serum
by GC-MS
23.4.2.4 Comments to the Procedure
23.4.3 Identification
23.4.3.1 Sample Preparation Procedure for Unknown
Screening by Mixed Mode Cation Exchange
23.4.3.2 Comments to the Procedure
23.4.3.3 GC-MS Procedure for Unknown Screening
23.4.3.4 Comments to the Procedure
23.4.3.5 LC-MS-MS Procedure for Unknown Screening
23.4.3.6 Comments to the Procedure
Index
xiii
464
465
466
466
466
466
467
467
467
468
468
468
468
468
469
471
471
472
472
472
473
473
475
475
477
Preface
This textbook, entitled “Introduction to Pharmaceutical Chemical Analysis”, is the first
textbook giving a systematic introduction to the chemical analysis of pharmaceutical
raw materials, finished pharmaceutical products, and drugs in biological fluids, as
carried out in the pharmaceutical laboratories worldwide. In addition to this, the
textbook teaches the fundamentals of all the major analytical techniques used in the
pharmaceutical laboratory and teaches the international pharmacopoeias and guidelines
of importance for the field. The textbook is primarily intended for the pharmacy student,
to teach the requirements in “analytical chemistry” for the 5-year pharmacy curriculum,
but the textbook is also intended for analytical chemists moving into the field of
pharmaceutical analysis.
The field of pharmaceutical analysis is very broad and challenging to define and limit, and
therefore we have made priority to some major areas of focus. First, the textbook has a major
focus on low-molecular-weight drug substances. This “low-molecular” focus was selected
to limit the size of the book, but also because we have a clear ambition of linking all the
discussions of the different chemical techniques and methods to the chemical properties of
the drug substances. We feel this is very important for a good understanding, and this
understanding is much easier to obtain for low-molecular drug substances than for
macromolecules. Thus, although macromolecules, like peptides and proteins, are also
used as drugs, they are not discussed in this textbook.
Second, this textbook has a major focus on pharmaceutical routine applications,
including how drug substances are analyzed as raw materials prior to pharmaceutical
production, how they are analyzed in finished pharmaceutical products, and how they are
analyzed in patient samples following administration. This “routine” focus was also
selected to limit the size of the book. Thus, applications of pharmaceutical analysis during
development of new drugs and during pharmaceutical research have not been discussed.
However, many of these applications are similar to the routine applications in terms of
fundamental understanding, and as long as the readers understand the routine applications,
they also have the best fundament to understand the more advanced applications.
Third, the textbook has a major focus on classical analytical techniques such as titration,
chromatography, electrophoresis, and spectroscopy. This “classical” focus was a natural
consequence of the “low-molecular” and “routine” focuses discussed above. Additionally,
we feel that discussing the most important techniques comprehensively is much more
valuable for the reader than mentioning all the techniques involved in pharmaceutical
analysis. In future revisions however, we may include more new analytical techniques as
they are gradually included as official methods in the international pharmacopoeias.
xvi
Preface
This textbook first gives a short introduction to the field of pharmacy, to the field of
pharmaceutical analysis, and to the regulations and guidelines relevant for the field
(Chapters 1 and 2). This is an important motivation for the reader, but is also a basis for
understanding the “landscape” of pharmaceutical analysis. Then, the textbook gives a short
chemistry course to make sure that the reader is at an appropriate level in terms of chemical
understanding (Chapter 3). This is very important, as we try to link every discussion later in
the book to chemical structures. The third part of the book (Chapters 4–20) describes all the
analytical techniques (tools). In this part of the textbook, we basically fill up the tool box to
be used in the final part. In the latter (Chapters 21–23), we describe how the different tools
(analytical techniques) are used for the analysis of pharmaceutical raw materials, for the
analysis of finished pharmaceutical products, and for the analysis of patient samples. Unlike
many other textbooks, we have no student problems. However, we have replaced the student
problems with many real examples, and for each example, we have given priority to
fundamental understanding of the chemistry and to calculations. Thus, in this textbook, you
learn how to calculate the concentration of a certain drug in your sample based on the
number displayed on your analytical instrument. Welcome to the challenging world of
pharmaceutical analysis!
Copenhagen/Oslo, April 2011
Steen Honore Hansen
University of Copenhagen
Knut Einar Rasmussen
University of Oslo
Stig Pedersen-Bjergaard
University of Oslo
University of Copenhagen
1
Introduction to Pharmaceutical Analysis
This chapter briefly reviews the life of medical products and the manufacture of medical
products according to international regulations and guidelines. Based on this review the
major areas and usage of pharmaceutical analysis are identified.
1.1 Applications and Definitions
The European Pharmacopeia defines a medical product as:
(a) Any substance or combination of substances presented as having properties for
treating or preventing disease in human beings and/or animals; or (b) any substance
or combination of substances that may be used in or administered to human beings
and/or animals with a view either to restoring, correcting or modifying physiological
functions by exerting a pharmacological, immunological or metabolic action, or to
making a medical diagnosis.
A medical product contains a substance that is pharmacologically active and that
substance is called the active ingredient (AI) or active pharmaceutical ingredient (API)
defined as follows:
Any substance intended to be used in the manufacture of a medicinal product and
that, when so used, becomes an active ingredient of the medicinal product. Such
substances are intended to furnish a pharmacological activity or other direct effect
in the diagnosis, cure, mitigation, treatment or prevention of disease, or to affect the
structure and function of the body.
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
2
Introduction to Pharmaceutical Chemical Analysis
An herbal medical product is:
A medicinal product, exclusively containing as active ingredients one or more herbal
drugs or one or more herbal drug preparations, or one or more such herbal drugs in
combination with one or more such herbal drug preparation.
Drug substances are administered very rare as the pure active substance. Typically the
active substance and excipients (auxiliary substances) are combined into dosage forms to
produce the final medical product. An excipient is:
Any constituent of a medicinal product that is not an active substance.
Adjuvants, stabilizers, antimicrobial preservatives, diluents, antioxidants, for example,
are excipients.
The dosage form can be, for example, a tablet or a capsule or syrup to be administered
orally, injections that are for parenteral administration into the body, or ointments for
topical administration. Figure 1.1 shows typical dosage forms.
Formulation is the process in which different chemical substances, including the active
ingredient and excipients are combined to produce a final medical product. It involves
developing a preparation of the drug that is both stable and acceptable to the patient. For
orally taken drugs this usually involves incorporating the drug and excipients in a solid
Syrup for oral
administration
Capsules
Tablets
Solution for injection
Figure 1.1 Different dosage forms
Introduction to Pharmaceutical Analysis
3
Table 1.1 Excipients of a paracetamol tablet and a paracetamol syrup
Content
Tablet (weight 285 mg)
Paracetamol
Hydroxypropyl cellulose
Maize starch
Talcum
Magnesium stearate
Syrup (volume 1 ml)
Paracetamol
Sorbitol
Glycerol
Polyvinylpyrrolidone
Saccharine sodium salt
Methylparabene
Ethylparabene
Propylparabene
Sodium metabisulfite
Citric acid
Sodium citrate
Strawberry aroma
Water
Amount (mg)
Function
250
Active ingredient
Binder
Disintegrant
Glidant
Lubricant
24
Active ingredient
Sweetener
Sweetener
Thickening agent
Sweetener
Preservative
Preservative
Preservative
Antioxidant
pH regulator
pH regulator
Flavoring agent
Solvent
dosage form such as a tablet or a capsule or a liquid dosage forms such as a syrup. The main
function of excipients is summarized as follows:
.
Ensure that the preparation has a shape and size that is easy to use for the patient;
Ensure that the active substance is optimally adsorbed in the patient;
. Ensure that the preparation has an acceptable shelf life;
. Ensure that the preparation does not have an unpleasant taste or odor;
. Ensure easy production.
.
There is a wide spectrum of different excipients, which varies widely from preparation
to preparation. To illustrate this, Table 1.1 shows the excipients of a tablet and syrup
which both contain paracetamol as the active ingredient. Paracetamol is both an analgesic
(an ¼ no, algesis ¼ pain) and a antipyretic (anti ¼ against, pyretos ¼ fever), which means
that it is used against pain and fever.
The tablets, which in this example, have a total weight of 285 mg contains 250 mg of
paracetamol (active ingredient), while the remaining 35 mg is made up of excipients.
The excipients are a disintegrating agent, a lubricant, a glidant and a binder. Binders,
lubricating and gliding agents are added to facilitate manufacture. A disintegrating agent
ensures rapid disintegration of the tablet in the stomach.
Paracetamol syrup, which contains 24 mg/ml of paracetamol, is composed mainly of
water. In addition, it is added sweetening and flavoring agents for better taste. Antimicrobial
preservatives and antioxidants are added to prevent bacterial growth and chemical degradation. In addition agents that increase the viscosity and stabilizes the pH are added.
Medical products may be divided into over the counter drugs (OTC), which may be
sold directly to the consumer in pharmacies and supermarkets without restrictions, and
4
Introduction to Pharmaceutical Chemical Analysis
prescription only medicine (POM) that must be prescribed by a licensed practitioner.
Medical products are predominantly produced by the pharmaceutical companies, only in
rare occasions are pharmaceutical products produced in hospitals and in pharmacies.
New products are often patented to give the developer exclusive right to produce them.
Those that are not patented or with expired patents, are called generic drugs since they can
be produced by other companies without restrictions or licenses from the patent holder.
According to the European Federation of Pharmaceutical Industries and Associations
(EFPIA), the pharmaceutical industry in Europe employed some 630 000 people, including
110 000 in research and development, in 2009. The trade surplus was Euro 55 200 million,
and Euro 26 000 million was spent on pharmaceutical research and development. The
retail value of the pharmaceutical market was Euro 215 000 million, which is just under
30% of the world market.
1.2 The Life of Medicines
Figure 1.2 outlines a typical industrial production of a pharmaceutical product.
Production starts by ordering the current active ingredient and the necessary starting
materials. In some cases, the company produces some of the ingredients, but most commonly
they are produced elsewhere by various industrial raw material suppliers. The raw materials
arrive in relatively large quantities (1–500 kg) and are typically packed in cardboard drums
or in plastic containers. Figure 1.3 shows an example of a received batch of raw material in
the photo gallery from a manufacturing facility.
Documentation and Control
Arrival of starting and packaging materials
Sampling of starting materials
Manufacturing
Filling
Labelling
Packaging
Documentation and control of finshed product
and product release
Figure 1.2 Illustration of the manufacturing of a medical product
Introduction to Pharmaceutical Analysis
5
Figure 1.3 Photo gallery of a manufacturing facility: arrival of raw materials, weighing,
sampling, tablet pressing, filling, labelling, and packaging. Reproduced with permission from
Fagbokforlaget
Figure 1.4 shows an outline of some areas found in a manufacturing facility.
Upon arrival the raw materials are registered in the manufacturer’s documentation
system, tagged with internal labels and stored in a separate area of the warehouse or in
a separate room where they are in quarantine until they are released for production. Samples
of the raw materials are collected and analyzed to ensure that the raw materials are of
6
Introduction to Pharmaceutical Chemical Analysis
Figure 1.3 (Continued)
satisfactory quality. This is the first of several important areas where pharmaceutical
analysis are vital. We focus further on this in Chapter 21. If the results are in accordance with
the specifications of the manufacturer the raw materials are labeled as released materials,
and transferred to the production facility. Production starts with weighing or measuring
the active ingredient and excipients in appropriate amounts for the subsequent production
(see Figure 1.3). Then, the raw materials are transferred to the manufacturing machinery.
Manufacture of tablets uses several types of equipment such as machinery for granulation,
drying and tablet pressing (see Figure 1.3). The manufacture of liquid dosage forms is
carried out in large tanks, while the production of ointments and creams are carried in
large pots with agitator and heating. When the product leaves the production site samples
Introduction to Pharmaceutical Analysis
Figure 1.3 (Continued)
7
8
Introduction to Pharmaceutical Chemical Analysis
for a comprehensive finished product control is collected. A number of analytical tests
are made, and this is another important field of pharmaceutical analysis which is discussed
in detail in Chapter 22. The products are in quarantine until the results of the testing show
compliance with specifications. The released product is filled in appropriate containers
(filling; see Figure 1.3), the containers are marked with labels (labeling; see Figure 1.3) and
the containers are packed in cardboard boxes (packaging; see Figure 1.3). Assessment of the
finished product embrace all relevant factors, including production conditions, results of inprocess testing, a review of manufacturing (including packaging), documentation, compliance with Finished Product Specifications and examination of the final finished pack.
As shown above, the industrial production of pharmaceuticals is a comprehensive process
that takes place over many different steps. Typically, production is a batch process, which
means that the products are made in limited batches. Each time a batch is produced a new
manufacturing process is started from the beginning with new starting materials. Between
each production of a given product, the equipment is often used for the production of other
products. Consequently the production facility must be cleaned thoroughly between each
batch to prevent the material from an earlier production contaminating other products
(cross contamination).
After leaving the manufacturer the products are sent to pharmaceutical wholesalers,
which provide for their further distribution to pharmacies, hospitals or other retailers
where they becomes available to the patients. Medicines have a broad scope of usage, and
are used against many types of illness and pain in various parts of the body.
At the start of medication, it is common to follow a standard treatment, but it is well
known that different patients may exhibit large variations in response. In such cases it is
important to adjust the dosage. One example is the treatment of hypertension. The dosage
may be reduced when the blood pressure is too low and the dosage may be increased
when the blood pressure is too high. For other types of treatment, such as depression,
psychosis and epilepsy, the measurement of effect is difficult; and in those cases therapeutic
drug monitoring (TDM) is advised. In TDM blood samples are collected and analyzed
to ensure that the drug level is appropriate. The analysis of drugs in biological fluids is
called bioanalysis. In addition to TDM, bioanalysis is crucial in drug development
programs, in forensic and toxicological analysis and in doping control testing in sports.
Bioanalysis is a third major area of pharmaceutical analysis, which is discussed in
Chapter 23.
1.3 The Quality of Medical Products
The purchaser of food and drink normally discovers that a product is associated with a
significant quality problem if has either an abnormal taste, unusual smell or a look that
seems abnormal. Medicines are, however, special. For example, there is no way patients can
decide whether a tablet contains the active ingredient, whether it is the correct dosage,
or whether any contaminants or degradation products are present. The patient is not in a
position to recognize that a medicine is incorrect or defective. The patient literally takes
medicines entirely on trust and is at the end of a chain of implicit trust which extends
back through administering, dispensing, prescribing and distributing, right back to those
Introduction to Pharmaceutical Analysis
Regulatory Affairs
9
Shipping and Receiving
Facilities Office
Manufacturing
Packaging and Labelling
QC- Batch release
QC - Analytical
Validation
Product
Development
Laboratory
Quality
Assurance
Figure 1.4 Outline of some areas found in a manufacturing facility
responsible for manufacture of the product. It is therefore mandatory that the pharmaceutical industry maintains the highest standards of quality in the development, manufacture and control of medical products.
Government heath authorities are naturally concerned about the quality of medicines,
and regulate the development, manufacture and marketing of medical products by a number
of laws and guidelines. These are discussed in Chapter 2. The regulations and guidelines
are to assure the safety, protection and well being of the consumer or patient. Two important
areas are:
.
.
Marketing authorization of medical products;
Manufacturing authorization of medical products.
Marketing authorization, also called a license, is required before any medicine can be used
to treat people. Only when the regulatory bodies are satisfied that the product works as it
should, and that it is acceptably safe, is it given a marketing authorization or product license.
The regulatory system also imposes rigorous standards on manufacturers. Manufacturing
authorization is required by all pharmaceutical manufacturers and ensures that only
authorized manufacturers manufacture all licensed products. Competent authorities regularly inspect the activities of the manufacturers and annually collect samples of marketed
medicines for assessment of quality. National Medical Control Agencies have the power to
10
Introduction to Pharmaceutical Chemical Analysis
withdraw a product from the market and to suspend production. These Agencies can also
prosecute a manufacturer if the law has been broken.
The holder of a manufacturing authorisation must manufacture medical products so as
to ensure that they are fit for their intended use. The products should comply with the
requirements of the marketing authorization and should not put patients at risk due to
inadequate safety, quality or efficacy. The attainment of quality is the responsibility of the
management, and it relies on a comprehensively designed and correctly implemented
system of Quality Assurance (QA) incorporating Good Manufacturing Practice (GMP) and
Quality Control (QC). The system should be fully documented. The basic concepts of QA,
GMP and QC are inter-related, as shown in Figure 1.5.
Quality Assurance is a wide-ranging concept, which covers all matters that influence
the quality of a product. It is the sum of all organized arrangements that are made to
ensure that medical products are of the quality required for their intended use. Quality
Assurance therefore incorporates GMP, which ensures that products are consistently
produced and controlled to the quality standards appropriate to their intended use and
as required by the Marketing Authorization. The basics of GMP are that all manufacturing
processes are clearly defined, systematically reviewed and shown to be capable of
consistently manufacturing products of the required quality. Quality Control is part of
GMP and is concerned with sampling, specifications and testing, and with the organization,
documentation and release procedures. Release procedures should ensure that the necessary
and relevant tests are actually carried out and that materials are not released for use, nor
QA
GMP
QC
Figure 1.5 Illustration of the QA/GMP/QC inter-relationship
Introduction to Pharmaceutical Analysis
11
products released for sale or supply, until their quality has been judged to be satisfactory.
The independence of the Quality Control Department from other departments is considered
fundamental. Quality Control is not confined to laboratory operations, but must be involved
in all decisions that may concern the quality of a product.
According to European regulations each batch of finished product must be certified by a
Qualified Person (QP) before being released for sale or supply.
Before certifying a batch the QP should ensure that at least the following requirements
have been met:
.
The batch and its manufacture comply with the provisions of the marketing authorization.
Manufacture has been carried out in accordance with GMP.
. The principal manufacturing and testing processes have been validated (validation is
defined as the documented act of demonstrating that processes will consistently lead to
the expected results).
. Any deviations or planned changes in production or quality control have been authorized
by the persons responsible in accordance with a defined system.
. All the necessary checks and tests have been performed.
. All necessary production and quality control documentation has been completed.
. The QP should in addition take into account any other factors of which he is aware
which are relevant for the quality of the batch.
.
Good documentation constitutes an essential part of the quality assurance system and
constitutes a vital part of batch release and certification by the QP. Clearly written documentation and standard operating procedures (SOP) prevent errors from spoken communication
and permit tracing of batch history. The documentation include:
.
Specifications that in detail describe the requirements that must be fulfilled prior to
quality evaluation;
. Manufacturing formulae, processing and packaging instructions;
. Procedures that give directions for performing operations such as cleaning, sampling
testing and equipment operation;
. Records providing a history of each batch or product.
The batch documentation shall be retained for at least one year after the expiry date of
the batches.
1.4 Summary
Government health authorities have regulated the development, manufacture and marketing of medical products by a number of laws and guidelines to assure the safety,
protection and well being of the patient. Market authorization is required before any
medical product can be marketed and only authorized manufacturers can produce
authorized products. Authorized manufacture is based on a correctly implemented system
of Quality Assurance incorporating Good Manufacturing Practice and Quality Control.
2
International Pharmacopoeias,
Regulations and Guidelines
This chapter reviews a number of laws, guidelines and regulations important in the
pharmaceutical production and thus also for the subject pharmaceutical analysis. The
manufacture of drugs is international and therefore this chapter focuses on international
affairs. But also national laws, guidelines, and regulations are important for manufacturers
in order to fulfil the requirements of the relevant authorities. All these laws, guidelines and
regulations are subject to regular updates and the latest edition should therefore always
be consulted.
2.1 Overview of Legislation
At the end of Chapter 1 it was mentioned that drug manufacturers have to fulfil requirements given in a number of laws, guidelines and regulations that will ensure high quality
of pharmaceutical preparations. Production and the related control of medicines are
often governed by the following two laws on a national basis:
.
.
Law on medicinal products (Medicines Act);
Law on pharmacy (Pharmacy Act).
The Medicines Act regulates, among other things, the requirements for medicines, clinical
trials (the testing of drugs on humans), the manufacturing process, import, wholesale sales,
retail sales and the advertising of medicines. This chapter focuses on what is relevant
in connection to drug development and production conducted in the pharmaceutical
industry and in pharmacies. We first discuss what is applicable for Proprietary Medicinal
Products manufactured industrially, while medicinal products produced in pharmacies
(stock production, small-scale production) are treated at the end of this chapter.
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
14
Introduction to Pharmaceutical Chemical Analysis
2.2 Legislation and Regulations for Industrial Production
A key point in the Medicines Act is that all medicinal products to be sold in a country must
have a marketing authorization granted by the local authorities (Ministry of Health). In the
EU a common legislation is available if a pharmaceutical company wants to register its
products in more than one EU country. This legislation is managed by the European
Medicines Agency (EMA). In the United States the Food and Drug Administration (FDA) is
the regulatory authority for the control of drugs.
The drug manufacturer or the importer of the drug is responsible for filing the application
for marketing authorization. The Ministry of Health through the State Medicines Agency
is responsible for evaluating the application based on the stated quality, safety and efficacy
of the drug. The market authorization is granted for five years and may thereafter be
renewed. The marketing authorization may be withdrawn before the expiration by the
Ministry of Health if:
.
The product is no longer considered to meet the requirements for quality, safety or efficacy.
The product does not have the specific qualitative or quantitative composition.
. The provisions that apply to pharmaceutical preparations have not been observed.
.
This means that a drug may be withdrawn from the market if there are problems associated
with its use, or if the composition of the preparation does not comply with the marketing
authorization. The Medicines Act contains no detailed information about what an application for marketing authorization shall contain, but some headlines are given in Table 2.1.
All analytical chemical control procedures and data for drug analysis that should be
performed to ensure the quality of the forthcoming product are to be described in Section 2
of the application for marketing authorization (chemical, pharmaceutical and biological
documentation). No further information about this is given in the pharmaceutical regulation,
but the applicant has to refer to the EU directives and guidelines (for Europe) or to FDA
(for USA) where further specification of what should be included can be found.
Table 2.1 Requirements for the content of an application for marketing authorization
Point
Description
1
2
3
4
5
6
7
8
9
10
11
12
13a
14a
Administrative information (name of applicant, name of manufacturer)
Name of the drug and its composition
Chemical, pharmaceutical, and biological documentation
Toxicological and pharmacological documentation
Clinical documentation
Expert reports
Proposals for advertisements
Proposed labeling preparations
Proposed leaflet
Proposals for prescription status
Documentation of the manufacturing authorization
Confirmation of fees paid to the Medicines Agency
A copy of the marketing authorization for the drug in other EEA countries
A copy of the advertisement and package leaflet approved in other EEA countries
a
Points 13 and 14 only apply within Europe.
International Pharmacopoeias, Regulations and Guidelines
15
Table 2.2 Excerpts from the content requirements of the application for marketing
authorization for medicinal product on chemical evidence. Only the topics most relevant to drug
analysis are included
Topic
Requirements
Control of active substance
(active pharmaceutical
ingredient; API)
Characteristics of the active substance and purity
requirements (specification)
Detailed description of the applicant’s chemical methods
for the confirmation of the identity of the active substance
Detailed description of the applicant’s chemical methods
for control of the purity
Detailed information on the active substance
Nomenclature of the active substance
Description of the active substance
Manufacturing method for the active substance
(chemical synthesis)
Quality control methods of the manufacturer of the
active substance
Known impurities in the active substance
Results of the chemical control of previously produced
batches of the active substance
Characteristics of the excipients and purity requirements
(specification)
Detailed description of the applicant’s methods for
confirmation of identity of the excipient
Detailed description of the applicant’s chemical methods
for control of the purity of the excipient
Information on excipients
Detailed description of the applicant’s chemical methods
for control of production mixtures
Detailed description of the applicant’s chemical methods
for the confirmation of the identity and determination of
the content of the active substance in the preparation
Detailed description of the applicant’s chemical methods
for the confirmation of the identity of a dye (excipient)
Detailed description of the applicant’s chemical methods
for the determination of the levels of antimicrobial
additives and preservatives
Documentation (validation) of the suitability of all the
chemical methods
Results of the chemical control of previously produced
batches of the product
Control of excipients
Control of production
preparations
Control of finished product
Table 2.2 lists an excerpt of the type of information that must be included on chemical
evidence based on applicable EU directives.
As you can see from Table 2.2, detailed information has to be filed about the methods to be
used to control the raw materials before production, about the methods to be used during
production and the methods to be used for control of the finished product. In addition, the
application must contain the results from the test analysis of the different trial production
batches, with documentation that the methods used have been tested and found suitable for
use (validation). The entire control scheme based on drug analysis for the forthcoming
16
Introduction to Pharmaceutical Chemical Analysis
preparations should therefore be documented in the application for marketing approval,
and this, together with the rest of the information shown in Table 2.1 forms the basis for
the authorities evaluation of the product application. The requirements for toxicological,
pharmacological and clinical documentation are different when applying for new drugs
or for preparations containing known active substances, and especially in the former case,
applications for marketing approval require very comprehensive documentation. If a
pharmaceutical manufacturer later wish to make changes in product composition or
changes of control methods it must be approved by the Ministry of Health at the National
Medicines Agency.
The Medicines Act states that any medicinal products can only be market if granted a
marketing authorization, and the law also states that anyone who manufactures drugs must
have a manufacturing authorization from the National Medicines Agency, and that all
premises and laboratories shall be in accordance with good manufacturing practice (GMP).
The word “manufacturing” in the Medicines act refers to production, packaging, repackaging, labeling, relabeling and release of drugs, as well as the necessary controls in connection
to these activities. Therefore analytical chemical work is included in the Medicines Act and in
GMP guidelines. The Medicines Act gives only little detail on the areas of manufacturing
authorization and good manufacturing practice, and more information about this should be
found in, for example, regulations on the manufacture and import of drugs. This regulation
specifies that a manufacturing authorization applies to certain premises at the manufacturing
site and only for certain drugs or drug types. This means that a pharmaceutical manufacturer
producing several different products must have a marketing authorization for each product,
while in principle only one manufacturing authorization is needed. The consequence of this
regulation is that the pharmaceutical analysis performed in conjunction with quality control
of pharmaceutical production has to be performed by a manufacturer with manufacturing
authorization for pharmaceuticals; and pharmaceutical analysis, like the rest of the
manufacturing process, is subject to good manufacturing practice (GMP). Good Laboratory
Practice (GLP) is another set of rules that should be consulted, and FDA, OECD and EU all
have rules, directives and so on in this field.
In the directive “Regulation on the manufacture and import of drugs” only few details
on what GMP means are given, but in this case, refer to the EU Directive. GMP is a
comprehensive regulatory framework, and the main content is presented in Table 2.3.
Table 2.3 The main elements of the GMP/GLP regulations
Paragraph
Description
1
Demands for a quality department and for quality control in all stages
of production
Requirements for staff
Requirements for premises and equipment
Documentation requirements
Requirements for production
Requirements for quality control
Requirements in connection with external contract work
Requirements for the withdrawal of products
Requirements for self-inspection
2
3
4
5
6
7
8
9
International Pharmacopoeias, Regulations and Guidelines
17
A brief discussion of the main headlines of the rules without going into further detail shall
be given.
Paragraphs 1, 2, 3, 4, 6 and 9 in Table 2.3 are of particular relevance for pharmaceutical
analysis related to production. Paragraph 1 of the GMP regulations requires all manufacturers
of drugs to have an overall system of quality assurance throughout the company. Manufacturers are also required to employ enough competent persons with the necessary
theoretical and practical skills (paragraph 2). The manufacturer is required to have an
organization chart, where each employee’s duties and responsibilities are defined, so that
there is no uncertainty about the duties and responsibilities. Under item 3, specified detailed
requirements for premises and equipment used during the manufacture of pharmaceuticals
are given. This ensures that the final products are of high quality and at the same time the
working environment of the staff is in accordance with official requirements. The latter is
important since drug production involves handling highly potent compounds. These rules
also apply to premises and personnel involved in drug analysis. Documentation is essential
throughout the manufacturing process including chemical analysis and quality control
(paragraph 4). Written documentation shall ensure that no misunderstandings occur during
manufacture. All information related to the production and control of individual production
sections shall be easily recovered for example at release of production batches, at inspections
and in the case of withdrawal of production batches. The requirement for documentation
applies to all departments in a manufacturing company, and is thus also an important part of
everyday life for the staff that performs drug analysis. Paragraph 6 of the GMP regulations
requires all companies that manufacture drugs to have a quality control (QC) department.
Furthermore, the company should have a quality assurance (QA) organization. This should
be sufficiently staffed with competent personnel and shall be independent of the other
departments in the company controlling the different steps in the manufacturing process.
Based on documentation from the various steps in manufacturing QA has the responsibility to
release products (discussed in Chapter 1). QA is thus also responsible for the drug analyses
performed (raw material control, production control, product control) and to interpret the
results from these. Manufacturers are under Clause 9 ordered to perform self-inspection,
which means that staff in the company at regular intervals must inspect the various
departments and ensure that they work as intended according to GMP.
The regulations on the manufacture and import of drugs oblige the manufacturer to follow
the manufacturing operations that are approved for marketing authorization. This means that
raw material control (Chapter 21), production control and product control (Chapter 22) must
always be performed for each production batch of the methods that are approved for marketing
authorization. In addition, the regulations require that each manufacturer appoints at least one
qualified person who is approved by the National Medicines Agency. Only qualified personnel
can assess the results from raw material control, production control and product control, and
together with the other documentation (see Chapter 1) release the production lots.
2.3 Life Time of Drugs and Drug Substances
Although some drugs (e.g., tablets) often have a stability that may last for many years
there is a maximum life time for all drugs and drug substances of five years. This is to
18
Introduction to Pharmaceutical Chemical Analysis
avoid any discussion on how long a time drugs and drug substances may be stored. However,
drug substances should always comply with the monographs in the pharmacopoeias or
similar standards.
Drug preparations in general should at the time of production at the most deviate 5%
from the declared content. On the basis of stability testing a larger deviation of the lower
limit of up to 10% within the life time of the drug is accepted. The life time is also named the
shelf-life of the drug.
2.4 Pharmacopoeias
Standards for finished pharmaceutical products and for raw materials used for such products
have been given in pharmacopoeias for many years. It all started in the seventeenth century
and many countries have their own pharmacopoeia. Pharmacopoeia Nordica 63 was the
first pharmacopoeia to be authorized in more than one country (the Nordic countries).
Now the European Pharmacopoeia (Ph.Eur.) – first published in 1969 – is authorized in
36 European countries (2010) and also the United States Pharmacopoeia is used in a
number of other countries.
The standards (also called monographs) in the pharmacopoeias are the official requirements for raw materials and finished products. Standards should be compiled as minimum
requirements and shall ensure that medicines are generally of high quality. In order to obtain
marketing approval, it is important that the manufacturer’s own requirements for the
product (see Table 2.2), which are called specifications, meet the government standards.
When using the monographs in the pharmacopoeias it is important also to be aware of the
provisions and requirements given in some of the general texts in the pharmacopoeias (e.g.,
test for residual solvents in not specified in each individual monograph but is a general
requirement for all raw materials). In addition, general monographs on dosage forms are
given in the Ph.Eur. The main contents of Ph.Eur. are given in Table 2.4.
Besides the European Pharmacopoeia (Figure 2.1) most European countries have a
national supplement with some local provisions. The British Pharmacopoeia includes
monographs on finished products similar to what can be found in USP.
A large number of standards for drug substances which are not still under patent are
included in Ph.Eur. and USP, and in these standards requirements for identity and purity
are given. Besides these requirements the pharmacopoeias also provide detailed procedures
for how to perform the identification and control of purity for each drug substance. In
the application for the marketing authorization, it is usual that the manufacturer makes
Table 2.4 Main content of the European Pharmacopoeia (Ph.Eur.)
Item
Description
1
2
3
4
Official purity requirements for raw materials (standards)
Official methods for control of raw materials
Official requirements for pharmaceutical formulations (dosage forms)
Official requirements for finished preparations (e.g., vaccines)
International Pharmacopoeias, Regulations and Guidelines
19
Figure 2.1 The European Pharmacopoeia (Ph.Eur.) 7th edn, 2011
reference to this information for the documentation of the chemical raw materials
(see Table 2.2). In case of production the producer has to prove that the requirements
for identity and purity are met, and normally he will refer to the methods given in
the pharmacopoeia.
There are also many active pharmaceutical ingredients (APIs) that is patent protected
by a single pharmaceutical company. Since there is normally only one company who has
the right to produce such a drug substance it is not necessary to describe it in the
pharmacopoeia, but the requirements (specifications) and procedures used by that company,
and approved by the Ministry of Health, are generally similar to those found in similar
monographs in the pharmacopoeias. As the patent rights of pharmaceutical raw materials
run out, they are normally included in the pharmacopoeias if they still are in use as
marketed medicines.
2.5 International Harmonization
Due to globalization and expansion in international trade there is a growing need to
develop global quality standards for medicines. As standards are a vital instrument for
registration, market surveillance and the free movement and trade of medicines among
as many countries as possible, harmonization among the world’s three major pharmacopoeias, the European Pharmacopoeia (Ph.Eur.), the Japanese Pharmacopoeia (JP)
and the United States Pharmacopoeia (USP), is an important task. This harmonization
process is now well under way in between the three pharmacopoeias but there is still long
way to go before the standardization of medicines becomes truely international.
20
Introduction to Pharmaceutical Chemical Analysis
Table 2.5 ICH guidelines of important relevance for the pharmaceutical analysis
Reference
Description
Q1A(R2)
Q2(R1)
Q2B
Q3A(R2)
Q3B(R2)
Q3C(R4)
Stability testing of new drug substances and products
Validation of analytical procedures: text and methodology
Validation of analytical procedures: methodology [in Q2(R1)]
Impurities in new drug substances
Impurities in new drug products
Impurities: guideline for residual solvents
2.5.1
International Conference on Harmonization
In 1990 a trilateral program, the International Conference on Harmonization (ICH), for the
harmonization of testing of medicines among the European Union, the United States and
Japan was set up. This program aims to reduce the overall cost of pharmaceutical research
and development worldwide by avoiding the duplication of work such as the preparation of
dossiers and studies, thus reducing the time required for innovative medicines to become
available. This conference takes places twice a year with the location of meetings rotating
among Europe, Japan and the United States. ICH has also published a number of guidelines
which have become important document especially for the pharmaceutical industry. The
most important ICH guidelines for analytical chemists are presented in Table 2.5.
2.5.2
Pharmacopoeial Discussion Group
Further work on harmonization between the three pharmacopoeias is carried out by the
Pharmacopoeial Discussion Group (PDG). The PDG considers proposals made by national
associations of manufacturers of pharmaceutical products and excipients in order to select
general methods of analysis and excipient monographs for addition to its work program. To
promote these exchanges and synergy, since 2001 the PDG has organized, upon request,
hearings for representatives of the pharmaceutical and excipient industries. Harmonizations
of monographs on active pharmaceutical ingredients have now also been initiated.
2.6 Legislation and Regulations for Pharmacy Production
So far we have focused on industrially produced drugs, but since pharmaceutical preparations are also produced on a small scale in many pharmacies, the regulations for these will
briefly be discussed. For drugs made in pharmacies The Pharmacy Act applies as the overall
legislation. The Pharmacy Act states that the Ministry of Health can provide manufacturing
authorization to the pharmacy. This means that, like the pharmaceutical industry, the
pharmacy must also have a manufacturing authorization in order to produce drugs. The
Pharmacies Act further stipulates that the manufacture shall be in accordance with good
manufacturing practice (GMP). Further reading of the regulations on the manufacture of
medicines in pharmacies given by the Ministry of Health explain that the raw materials and
finished drugs shall meet the requirements of the Ph.Eur. Normally the pharmacy buys raw
International Pharmacopoeias, Regulations and Guidelines
21
materials which have been tested for identity and purity according to Ph.Eur. or another
official standard by the supplier, and therefore the pharmacy only has to secure identity.
Smal-scale preparation of products having no marketing authorization does not require a
final testing for the content by drug analysis. The pharmacy is obliged to otherwise ensure
that the quality is consistent with the Ph.Eur., and this is executed by working according to
GMP. If the pharmacy produces preparations on a larger scale, which is called the stock
production, the pharmacy is required to perform final testing of the content.
2.7 Summary
Pharmaceutical research related to quality control of pharmaceutical production (raw
material control, production control, product control) is intended to verify that the current
specifications are met. The area of drug substances, drug development and drug production
is heavily controlled by a number of laws, regulations and guidelines. All chemical methods
for control of a preparation shall be documented in the marketing authorization. The
government standards are given in pharmacopoeias, and manufacturers are obliged to
follow these standards or to prepare their own specifications of similar or even better quality.
Drug manufacturers should have a manufacturing authorization, and production and
inspection shall be performed in accordance with good manufacturing practice (GMP).
3
Fundamental Chemical Properties,
Buffers and pH
Chemical analysis is an important part of the quality assessment of drugs. A deeper
understanding of the analytical method requires knowledge about both the analytical
technique and the chemical properties of the analytes. Therefore a basic knowledge of a
number of physicochemical properties of molecules is needed to be able to understand and
further develop analytical chemical methods. For example, knowledge of spectroscopic
principles and techniques is needed when the choice of the detection technique for a
given separation method is made; and knowledge of pH and pKa values is important for
the design of many sample preparation techniques. Spectroscopic techniques are treated
in separate chapters, and this chapter discusses some important physicochemical and
chemical properties of drug substances and focuses on how to utilize the information in an
analytical chemical context.
3.1 pH and pKa
pH is an expression for the acidity or alkalinity of an aqueous solution and the pH
concept is extremely important, having great influence on living organisms as well as in
analytical chemistry.
Water can react with itself to form a hydronium ion and a hydroxide ion:
2H2 O a H3 Oþ þ OH
ð3:1Þ
This is called autoprotolysis as the water acts as an acid as well as a base. The autoprotolysis
constant is:
Kw ¼
½H3 Oþ ½OH
½H2 O2
¼ 1014
ð3:2Þ
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
24
Introduction to Pharmaceutical Chemical Analysis
indicating that only a very small amount of water is ionized. The concentration of the two
ions in pure water is therefore 107 M of each ion.
pH is defined as the negative logarithm to the activity, aHþ, or the concentration of the
hydrogen ions (equivalent to the hydronium ions), [Hþ].
pH ¼ logðaHþÞ logð½HþÞ
ð3:3Þ
Strong acids and strong bases are fully ionized in dilute aqueous solution and the activity and
their concentrations of [Hþ] therefore can be considered to be identical.
Weak acids and weak bases are not completely ionized in aqueous solution and are
therefore in equilibrium with the unionized acid or base. When we ignore the weak
autoprotolysis of water we get the following general equation for a weak acid:
HA a Hþ þ A ) K a ¼
½Hþ ½A
½HA
ð3:4Þ
When acid in form of Hþ is added to such a system, the Hþ will partly be removed and form
HA, and if a base is added it will partly be neutralized by Hþ and more HAwill dissociate. pH
will thus be maintained in the solution. A system like this is called a buffer system and the
purpose of a buffer is to maintain the pH in the solution.
As:
pKa ¼ logKa
ð3:5Þ
it is obvious that the highest buffer capacity is achieved at a pH value close to the pKa
value of the buffer substance. This brings us to a most useful equation called the
Henderson–Hasselbalch equation:
pH ¼ pKa þ log
½A
½HA
ð3:6Þ
At pH ¼ pKa equal concentrations of the acid and corresponding base are present. If the ratio
between HA/A becomes 9 : 1 (only 10% base) pH decreases by one unit; and if the ratio
becomes 99 : 1 (1% base) the pH value decreases by two units.
Equivalent estimations can be performed when increasing the base content.
It is convenient to have an knowledge of pKa values of a number of functional groups
as presented in Table 3.1.
Table 3.1 Typical pKa values of functional groups
Functional group
pKa
Comment (depending on
chemical structure)
R-COOH, carboxylic acid
R-NH2; R1,R2,NH and R1,R2,R3,N,
aliphatic amines
Ar-OH, phenols
R-OH, alcohols
4–5
8–10
Can be lower (more acidic)a
Can be lower (less basic)a
8–10
> 14
Can be lower (more acidic)a
Can be considered as neutral
substances
R-SO2OH, sulfonic acid
Approx. 1
a
This depends on other groups in the molecule.
Fundamental Chemical Properties, Buffers and pH
25
Table 3.2 Energy in bonds or of intermolecular forces
Type of bond or
intermolecular force
Example of interacting
molecules
Covalent
Ionic
Hydrogen bond
Dipole–dipole
Dipole–induced dipole
Dispersion/van der Waal
RH2C–CH2R
R4NþOOC-R
H3COH.....HOH
C6H5Cl/H3CCN
H3CCN/C6H6
C6H6/C6H14
Energy
kJ mol1
kcal mol1
400–1200
200–800
20–50
12–40
10–25
5–20
100–300
50–200
5–12
3–10
2–6
1–5
The pKa value for bases refers to the protonated form of the bases. However, the basicity
of bases may also be expressed equivalent to the pKa of acids. In that case the term pKb is
used and:
pKa þ pKb ¼ 14
ð3:7Þ
3.2 Partition
The partitioning of substances between immiscible phases (gas–liquid, gas–solid, liquid–liquid
or liquid–solid) is of major importance for a number of analytical chemical techniques.
Sample molecules introduced into the two-phase system will be exposed to a number of
interactions (diffusion, collisions, dipole–dipole interactions, hydrogen bonding, electrostatic interactions, etc., as illustrated in Table 3.2) in the two phases. The interactions taking
place are dependent on the physical and chemical nature of the analytes as well as of the
mobile and stationary phases and may result in different partition of analytes between the
two phases.
The partition between phases is also influenced by pH, and thus a thorough knowledge
of the pH concept including pKa as well as the distribution constant will ease the development of analytical methods (e.g., chromatographic methods). The distribution constant is
dependent on the nature of the two phases as well as the temperature. If we want to alter
the partition between the two phases, we must change one of these variables. The
equilibrium distribution for a substance A is given by the partition ratio, also called the
distribution constant, KD:
KD ¼
½Aorg
½Aaq
ð3:8Þ
where [A]org is the concentration of compound in the organic phase and [A]aq is the
concentration of compound in the water phase.
The distribution constant is a constant relating to a specific molecular species, but often the
molecules can be present as different species, for example, by dissociating in the aqueous phase:
HAþ H2 O a Aþ H3 Oþ
ð3:9Þ
26
Introduction to Pharmaceutical Chemical Analysis
or dimerizing in the organic phase:
2HA a ðHAÞ2
ð3:10Þ
These equilibria are very fast, and it is therefore appropriate to look at the total distribution of all
the species of a compound between the two phases:
DC ¼
½HAorg þ ½Aorg þ ½ðHAÞ2 org
½HAaq þ ½A aq þ ½ðHAÞ2 aq
¼
½HAtotal org
½HAtotal aq
ð3:11Þ
The concentration distribution ratio, DC, between the two phases can also be expressed as the
mass distribution ratio, Dm, by multiplying the concentrations with the matching phase
volumes:
Dm ¼
½HAtotal org Vorg
½HAtotal aq Vaq
¼
½ðamount of HAÞtotal org
½ðamount of HAÞtotal aq
½ðamount of HAÞtotal stat
½ðamount of HAÞtotal mob
ð3:12Þ
Where Vorg and Vaq refer the volumes of organic and water phases, respectively. The
subscript terms stat and mob refer to the stationary phase and the mobile phase used
in chromatography.
The fundamentals of partition are outlined in Chapter 18. The greater the partition coefficient the higher the affinity towards an organic phase will be. In the case of distribution to a
solid phase the partition can be governed by other characteristics than partition coefficients.
The partitioning of analytes in a system where one phase is a gas necessitates that the
analytes can enter the gas phase.
Discussions on the extraction and partition of compounds therefore most often refer to
liquid/liquid systems. Partition ratios are estimated using distribution between n-octanol
and water. If the compound can be ionized, the ionized form will have a much stronger
affinity towards the aqueous phase as water molecules will solvate the ions. The distribution of an ionizable compound therefore very much depends on the pH of the aqueous
phase. From the Henderson–Hasselbalch equation given above the following equations
can be derived:
DC
1þ 10pH pKa
DC
For bases : Dapp ¼
1þ 10pKa pH
For acids : Dapp ¼
ð3:13Þ
ð3:14Þ
If the distribution ratio, DC, and the pKa value are known for a compound, the apparent
distribution ratio, Dapp, at a given pH can be calculated.
Parameters such as the partition ratios in octanol/water are available as the so-called
log P values, and the distribution ratio of compounds between octanol and water at different
pH values in the water phase is tabulated as log D values. Computer programs can also be
used for estimation of pKa values, log P values, log D values and solubility of compounds.
The actual values of each parameter can vary when consulting different literature references,
and this is most often due to differences in the methods used for analysis. This is particularly
true for log P, log D and solubility data. In an analytical chemical context such parameters
should primarily be used as a guide.
Fundamental Chemical Properties, Buffers and pH
COOH pKa = 3.0
log P = 3.72
OH
pKa = 13.7
27
O
OH
pKa = 4.43
log P = 2.0
Ibuprofen
Salicylic acid
Figure 3.1 Chemical structures of ibuprofen and salicylic acid with log P and pKa values
Liquid/liquid extraction is often used in sample preparation. It is therefore of interest to
determine the fraction of analytes extracted under given conditions. This is given by the
general formula:
2
3n
1
5
En ¼ 1 4
1þ DC VV21
ð3:15Þ
where En is the extracted fraction, DC is the distribution ratio between the two phases V2
and V1 and n is the number of extractions. V1 is the phase that originally contains the
analyte and V2 is the phase to which the analytes is extracted.
Example 3.1: Ibuprofen (Figure 3.1) has a log P value of 3.72 and a pKa value of 4.43. If
10 ml of a sample solution of ibuprofen at pH 6.0 is to be extracted to 30 ml of an organic
solution, how much will be extracted?
A log P value of 3.72 corresponds to a KD of 5248. At pH 6.0 the apparent distribution
ratio will be 5248/1 þ 37 ¼ 138 (using Equation 3.13). Calculating the extracted fraction
gives 0.9997 or 99.97%. Doing the same extraction at pH 7.0 results in an extraction of
97.7%. Try to do this calculation yourself.
Example 3.2: Salicylic acid has a log P of 2.0 (corresponding to a KD of 100) and
pKa values of 3.0 and 13.7. Performing similar calculations as in the above example,
it can be shown that only 19% of the salicylic acid is extracted into 30 ml of organic
phase at pH 6.0. Lowering the pH to 5.0 gives an apparent Dc of about 1 and thus an
extraction of 75%.
A question could be whether multiple extractions using the same total amount of
organic solvent would improve the extraction yield. Consider using three extractions of
only10 ml each of organic solvent. Calculations using Equation 3.15 show that the total
extraction in the combined 30 ml will be 87.5% compared to the 75% obtained in only
one single 30-ml extraction. Multiple extractions are more efficient, but in the case of
salicylic acid it is necessary to perform the extraction at a lower pH value if a quantitative
extraction is needed.
Similar extraction calculations can be performed for bases using Equations 6.14 and 6.15.
It is obvious that quantitative extractions from an aqueous solution into an organic
phase are more easily achieved if extraction is performed when the analytes are not ionized.
Thus extraction of carboxylic acid should take place at low pH (pKa 2 or 3 pH units) and
bases at high pH (pKa þ 2 or 3 pH units).
28
Introduction to Pharmaceutical Chemical Analysis
3.3 Stereochemistry
The pharmacological activity of drug substances is very dependent on their physicochemical behavior, often discussed as their ADME properties (adsorption, distribution,
metabolism, excretion). The above-mentioned physicochemical parameters are of vital
importance, but also the stereochemistry of the drug substances is of importance as this
will affect the ADME properties. Describing the stereochemistry of a drug substance is
to visualize the spatial orientation of its components in space. Biological systems including
the human body contain large biomolecules which are constructed from building blocks
with unique stereochemistry. Biological systems are therefore able to distinguish between
isomers which only differ in their spatial configuration. Such isomers may therefore also
have different biological effects.
Figure 3.2 shows how isomers can be divided into several groups. Constitutional isomers
are of course different compounds with different chemical structures.
The diastereomers grouped under the stereoisomers contain compounds where the
isomers have different physicochemical characteristics with different melting points,
partition ratios and so on. These isomers are therefore easy to separate in chromatographic
systems. Cis–trans isomers belong to this group and a number of drug substances can be
found in this group. Two examples are given in Figure 3.3.
Isomerization at the double bond is often mediated by light and the compound should
therefore be protected from light exposure.
Enantiomers constitute a special group of stereoisomers. The two enantiomers that
constitute a pair contain a chiral center and are mirror images of each other. A chiral center
is an atom connected to three (S and P atoms) or four (N and C atoms) different ligands.
The most abundant chiral center is where a carbon atom is connected to four different
groups, but also nitrogen, phosphor and sulfur can be chiral centers. A compound containing
one or more chiral centers is able to rotate plane-polarized light either left or right. This is
Figure 3.2 Classification of isomers
Fundamental Chemical Properties, Buffers and pH
S
29
OH
HO
Cl
N
N
Clopenthixol
OH
OH
Resveratrol
Figure 3.3 The chemical structure of cis-clopenthixol and trans-resveratrol
denoted ( ) or (þ), respectively. However, this is not an unambiguous way to describe the
configuration of the chiral center as the direction and size of the rotation is dependent on
the solvent used for sample preparation. In older literature the terms d (dexter) and l (laevo)
were used to denote (þ) and ( ), respectively, but also the small capital letters D and L
have been used where reference was made to the configuration of glyceraldehyde. To give
an unambiguous description of the configuration of the chiral center the R/S nomenclature
has to be used. This nomenclature gives the absolute configuration of the position of groups
connected to the chiral atom and this nomenclature should always be used.
A pair of enantiomers has besides the rotation of the plane-polarized light identical
physicochemical characteristics. However, when they enter a chiral environment (e.g., a
biological system) they may behave differently. A large number of drug substances are
chiral, and for enantiomers it is often observed that one enantiomer has the beneficial
pharmacological effect while the other is either inactive or even gives rise to unwanted side
effects. It is therefore important to be able to control the purity of an enantiomeric drug
substance for content of the unwanted enantiomer.
Measuring the rotation of the plane-polarized light is one way to do this, but the technique
is not sensitive and a low percentage of the impurity cannot be detected. A chiral environment
is needed in order to separate the two enantiomers. This can be done by chromatography or
electrophoresis by introducing chirality into the system. This is described in Section 11.5.
3.4 Stability Testing
Drug substances and drug products should be stable or only degrade to a small extend during
their lifetime. It is therefore necessary to perform stability testing to obtain knowledge of
possible degradation processes and in this way establish the shelf-life of products. The
international ICH guidelines as well as guidelines from FDA in the United States describe
how to perform such studies. Many drug substances are fairly stable and under proper
storage are stable for at least five years, which is the normal authorized lifetime of a drug
substance or a drug. Also a number of dry drug formulations are very stable, but when it
comes to liquid preparations long time stability cannot be expected. Some drug substances
are susceptible to hydrolysis and/or oxidation which more readily take place in solution.
Reaction kinetics is used to calculate the shelf-life of pharmaceutical products. Depending
30
Introduction to Pharmaceutical Chemical Analysis
on the chemical reaction taking place the reaction kinetics can be divided into zero-, first-,
second-, or third-order reactions.
In zero-order reactions the reaction rate is independent of the analyte concentration. A
rate constant of 0.02 mmol/h corresponds to a degradation of 0.02 mmol/h and 0.048 mmol/
day. A solution containing 1 mol of a drug substance in this case degrades to 90% of its
original concentration (10 mmol of total degradation) within about 208 days. This gives the
product a shelf life of 0.5 year.
In most cases reaction kinetics are considered to be of first order or are approximated to a
first-order reaction, also denoted pseudo first order. In this case the reaction rate is dependent
on the analyte concentration [A], and the unit of the rate constant, k, is time1 (e.g., h1 or s1):
d½A
¼ k½A
dt
The degradation process by time can be described as:
ðx
ðt
dx
dx
¼ kða xÞ a
¼ kdt
dt
0 ða xÞ
0
ð3:16Þ
ð3:17Þ
By rearrangement and integration and also taking the amount of degraded product, x,
into consideration the following equation can be derived:
t¼
1
a
ln
k a x
ð3:18Þ
And from this the half life of the substance can be calculated:
t0:5 ¼
1
a
1
ln
¼ ln 2
1
k a2a k
ð3:19Þ
The hydrolysis of aspirin (acetylsalicylic acid) is considered as a pseudo first-order reaction
and at pH 7.4 and 25 C the rate constant is about 1.4 102 h1. Using Equation 3.19 gives a
half life of 49 h and a 10% degradation takes place within 7.5 h.
It is therefore not possible to store liquid preparations of aspirin as the shelf life would be
only a few hours. Due to the relative fast degradation it is also important to consider the
stability of the prepared sample when performing analysis of aspirin tablets. In order not to
bias the obtained quantitative analysis data, the extracted tablet solution should be analyzed
as quickly as possible. Degradation to 0.5 or 1.0% takes place within 0.36 or 0.7 h,
corresponding to about 20 or 40 min, respectively. Longer storage of the sample solution
increases the bias of the analytical data.
3.5 Summary
The examples show that many drug substances have poor water solubility, and that
substances with high log P values are less water soluble than compounds with a lower
log P. The solubility is often better in organic solvents. Many organic bases are available
as hydrochlorides, sulfates or phosphates which generally are soluble in water. Water
Fundamental Chemical Properties, Buffers and pH
31
solubility may be increased up to several thousand times if the pH is changed in order to
ionize functional groups.
In general, substances with high log P values are easily extracted from aqueous solutions
using an organic solvent. The higher the log P, the more effective will be the extraction.
Extraction may be performed in one or several steps. Substances with medium log P values
should generally be extracted using multiple extractions in order to have a high recovery.
The water phase must have a pH which suppresses ionization. Substances with high log P
values will also be retarded well in reversed phase chromatography.
4
Fundamentals of Pharmaceutical
Analysis
This chapter discusses the basics of pharmaceutical analysis, including the different types of
calculations related to pharmaceutical analysis. The chapter also includes a review of
simple laboratory equipment, how to make solutions and dilutions, how to calibrate
analytical methods, and how to use simple statistics on the analytical data. The chapter
concludes with a list of important terms and concepts in pharmaceutical analysis. It is
important that you read carefully through this chapter before proceeding to the subsequent
chapters. Important terms should be learned.
4.1 What is a Pharmaceutical (Chemical) Analysis?
A pharmaceutical analysis is intended to either identify or quantify one or more
substances in a given sample of pharmaceutical interest. In pharmaceutical analysis,
the substance or substances of interest are normally active pharmaceutical compounds,
pharmaceutical excipients, contaminants, or drug metabolites. A substance to be identified or quantified is called the analyte. The samples in pharmaceutical analysis are
typically pharmaceutical raw materials, finished pharmaceutical products, or biological
samples like human blood or urine containing one or more drug substances. The samples
consist of one or several analytes, and a sample matrix which is the rest of the sample.
Identification is intended to confirm the identity of the analytes. Identification can also be
referred to as qualitative analysis. A quantitative analysis is intended to measure the exact
concentration or the exact amount of the analyte in a given sample. A quantitative analysis
is also termed as a determination. As an example, paracetamol tablets containing 500 mg
paracetamol per tablet have to be controlled prior to release from production. This is
accomplished by pharmaceutical analysis. Paracetamol is the analyte, whereas the rest of
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
34
Introduction to Pharmaceutical Chemical Analysis
Sampling
Sample storage
Sample preparation
Analytical measurement
Figure 4.1 Different steps in a typical procedure in pharmaceutical analysis
the tablet, consisting of different pharmaceutical excipients, is the sample matrix.
Identification of paracetamol in the tablets is performed to make sure that the
tablets contain the correct active pharmaceutical ingredient, whereas a quantitative
analysis is performed to measure the content of paracetamol and to check that this result
is exactly or close to 500 mg per tablet. In the latter case, a determination of paracetamol
is performed.
Procedures for pharmaceutical analysis are often complicated and consist of several
steps, as illustrated in Figure 4.1.
First, a sampling is performed, where the required number of samples are taken. During
sampling, it is essential that samples are taken in a representative manner, to give a correct
picture of the case under investigation. Exemplified with the 500-mg paracetamol tablets
discussed above, finished tablets have to be sampled in a systematic way during the entire
time scale of production to give an average of the total production. Sampling is beyond the
scope of this book and is not discussed further here. However, sampling is subject to much
focus in the laboratories involved in pharmaceutical analysis. Often, the samples must be
stored until further analysis. Sample storage is also a very important point, and the reason for
this is to avoid compositional changes of the sample during storage. If compositional
changes occur, the final analytical result will not reflect the original composition of the
sample. To protect samples, they are frequently stored at low temperature and protected
from light, as in a refrigerator or a freezer. This is especially important for liquid samples
where degradation chemistry likely occurs. For the example with 500-mg paracetamol
tablets, sample storage is not a critical issue as the tablets are relatively stable at room
temperature. Sample storage is also beyond the scope of this book and is not discussed
further here.
After sampling and storage, samples are normally pretreated in some way, and this is
called sample preparation. Sample preparation can be very simple or quite complicated,
depending on the sample. Sample preparation is discussed in detail later in this book.
Sample preparation ensures: (i) the sample can be analyzed in a subsequent step, (ii) the
sample is compatible with the final analytical method or instrument, (iii) the analyte is
present in a sufficient amount to be detected, and (iv) substances in the sample matrix that
can cause problems or interferences are removed. In the example with 500-mg paracetamol tablets, the sample preparation normally includes pulverization of the tablets,
dissolution of the tablets, and filtration of material that has not dissolved. Finally, the
analytical measurement is performed, where the analyte or analytes are identified and
quantified. In this textbook, focus will principally be directed on analytical measurements
by titration, spectroscopy, chromatography, and electrophoresis, as discussed in subsequent chapters. In the example with 500-mg paracetamol tablets, spectroscopy is typically
used to identify paracetamol and to measure the quantity of paracetamol in the tablets.
After the analysis, the measurements are processed; the results are calculated and
presented in an analytical report.
Fundamentals of Pharmaceutical Analysis
35
4.2 How to Specify Quantities and Concentrations?
In most cases, a quantitative pharmaceutical analysis is performed when the analyte is
present in a solution. A solution is a homogeneous mixture of two or more substances. A
minor species in a solution is called a solute and the analyte is an example of a solute. The
major species in a solution is the solvent. The amount of analyte (solute) is normally
expressed as the concentration. Concentration means the amount of solute per volume unit
of solution. In some contexts, the term molarity is used to express the concentration, which is
abbreviated M. The molarity of a certain solute is defined as follows:
Molarity ¼ number of moles of solute per liter of solution
ð4:1Þ
An example of how to calculate the molarity during the preparation of a solution with a
known concentration of solute is given in Box 4.1.
When the concentrations are low, it is impractical to use M. In such cases, either
millimolar (mM), micromolar (mM), or nanomolar (nM) are used as defined in the following
way:
1M ¼ 103 mM ¼ 106 mm ¼ 109 nM
ð4:2Þ
An example of how to convert from M to mM is shown in Box 4.2.
As an alternative to molarity, it is common to use concentrations expressed as mass per
volume unit. Either milligrams per milliliter (mg/ml), micrograms per milliliter (mg/ml),
nanograms per milliliter (ng/ml), or picograms per milliliter (pg/ml) can be used, which are
defined as follows:
1 g=ml ¼ 103 mg=ml ¼ 106 mg=ml ¼ 109 ng=ml ¼ 1012 pg=ml
ð4:3Þ
An example of how to convert from mg/ml to mg/ml is shown in Box 4.3, and an example of
how to convert from mg/ml to M is shown in Box 4.4.
Box 4.1 Calculation of molarity
0.100 g paracetamol (molar mass ¼ 151.2 g/mol) is dissolved in water and the total
volume is adjusted to 500.0 ml. The molarity of paracetamol is calculated to:
0:100 g=151:2 g=mol
¼ 1:32 103 M
0:5000 l
Box 4.2 Conversion from M to mM
One sample has a content of 1.62 105 M paracetamol. This corresponds to the
following concentration in mM:
1:62 105 M 106 mM=M ¼ 16:2 mM
36
Introduction to Pharmaceutical Chemical Analysis
Box 4.3 Conversion from mg/ml to mg/ml
One sample has a content of 0.0125 mg/ml paracetamol. This corresponds to the
following concentrations in mg/ml:
0:0125 mg=ml 103 mg=mg ¼ 12:5 mg=ml
Box 4.4 Conversion from mg/ml to M
One sample has a content of 12.5 mg/ml of paracetamol (Molar mass ¼ 151.2 g/mol).
This corresponds to the following concentration in M:
12:5 mg=ml ¼ 12:5 g=l )
12:5 g=l
¼ 8:27 102 M
151:2 g=mol
Concentrations can also be expressed as a percentage. The following definitions are used:
mass of solute
100%
mass of solution
ð4:4Þ
volume of solute
100%
volume of solution
ð4:5Þ
% weight ¼ %ðw=wÞ ¼
% volume ¼ %ðv=vÞ ¼
% weight=volume ¼ %ðw=vÞ ¼
mass of solute
100%
volume of solution
ð4:6Þ
When expressing concentration as a percentage, it is highly important to specify whether %
weight, % volume, or % weight/volume has been used. % Weight is used to express the
concentration of solids or liquids either in solid samples or liquid samples. % Volume,
however, is used to express the concentration of liquids in liquid samples (or gases in
gaseous samples). % Weight/volume is used to specify the concentration of solids or liquids
in a solution. Boxes 4.5 and 4.6 show examples of how to calculate concentrations using the
terms % weight and % weight/volume, respectively.
At very low concentrations, it may be convenient to use the terms parts per million (ppm)
or parts per billion (ppb) instead of % weight. These are defined as follows:
Box 4.5 Calculation of % weight
5.01 g paracetamol is dissolved in 200.0 ml of ethanol, and the total weight of the
solution is determined to 162.81 g. The concentration of paracetamol in % weight is:
% ðw=wÞ ¼
5:01 g
100% ¼ 3:08% ðw=wÞ
162:81 g
Fundamentals of Pharmaceutical Analysis
37
Box 4.6 Calculation of % weight volume
5.01 g paracetamol is dissolved in ethanol and the volume of the solution adjusted to
200.0 ml. The concentration of paracetamol in % weight volume is:
% ðw=vÞ ¼
5:01 g
100% ¼ 2:51% ðw=vÞ
200:0 ml
Box 4.7 Calculation of ppm
0.106 g ethanol dissolved in 999.911 g water. The concentration of ethanol in ppm is:
ppm ¼
0:106 g
106 ¼ 106 ppm
0:106 g þ 999:911 g
ppm ¼
mass of solute
106
mass of sample
ð4:7Þ
ppb ¼
mass of solute
109
mass of sample
ð4:8Þ
Box 4.7 shows an example of how to calculate concentration in terms of ppm.
For the analysis of pharmaceutical raw materials and preparations, the analyte concentrations are relatively high, and normally mg/ml is used to express concentrations. However,
for the analysis of drugs in biological samples, the analyte concentrations are normally
much lower, and here it is customary to express the concentrations in ng/ml or nM. A drug
concentration of 1 ng/ml, which is normal for several drugs in biological samples, is an
extremely low concentration, as visualized in Box 4.8. Box 4.9 shows an example how to
convert concentrations from ng/ml to nM.
4.3 Basic Laboratory Equipment
4.3.1
The Analytical Balance
Quantitative pharmaceutical analysis is always based on accurate weighing of the substance
or sample to be analyzed. Such weighing should be performed with an analytical balance to
get as high accuracy as possible. The analytical balance is therefore a fundamental
instrument in the pharmaceutical laboratory. Many laboratories have laboratory balances
available, but these do not provide sufficiently high accuracy for analytical purposes. An
analytical balance is shown in Figure 4.2. Analytical balances are equipped with a digital
display with direct recording of the mass. The most common analytical balances have a
weighing capacity up to 100 or 200 g. The sensitivity of the balance is defined as the smallest
increment of mass that can be measured, and the most common analytical balances have
38
Introduction to Pharmaceutical Chemical Analysis
Box 4.8 Illustration of typical drug concentration in human blood
A patient sample contains 1.0 ng/ml of fluphenazine. To visualize how little substance
this is per volume unit, one can imagine that a sugar cube is dissolved in water. How
much water is required to give a concentration of 1 ng/ml?
The weight of a sugar cube is about 2.3 g. The following amounts of water (V) must be
used to give a concentration of 1.0 ng/ml:
1:0 ng=ml ¼ 1:0 109 g=ml ¼ 1:0 106 g=l ¼ 2:3 g=V
2:3 g
V¼
¼ 2:3 106 l
1:0 106 g=l
2.3 106 l equals about the amount of water in a swimming pool which is 2.5 m deep,
20 m in width and has a length of 50 m. A drug concentration of 1 ng/ml is equivalent to
the concentration you get if you solve a sugar cube in a large swimming pool! As you will
learn in this book, such low concentrations can be measured with our analytical
techniques!
With a concentration of 1.0 ng/ml, how many molecules are present per ml? Assume
the molar mass is 300 g/mol. In 1.0 ml, 1.0 ng drug is present, which is equal to
1.0 109 g, and which corresponds to the following number of moles:
n¼
1:0 109 g
¼ 3:3 1012 mol
300 g=mol
This number of moles corresponds to the following number of molecules, by multiplication by Avogadro’s number:
Number of molecules ¼ 3:3 1012 mol 6:022 1023 molecules=mol
¼ 2:2 1012 molecules
As you can see, although dealing with very low concentrations, there are still a lot of
molecules!
sensitivities in the range 0.01–0.1 mg. The analytical balance should be located on a heavy
table, such as a marble slab, to minimize vibrations. This is important to ensure stable
readings. Analytical balances have adjustable feet and a bubble meter that allow you to keep
the balance level.
Box 4.9 Conversion from ng/ml to nM
A patient sample was found to contain 1.0 ng/ml fluphenazine. This substance has a
molar mass of 437.58 g/mol. The concentration is converted to nM in the following way:
1:0 ng=ml ¼ 1:0 mg=l ¼ 1:0 106 g=l )
1:0 106 g=l
¼ 2:3 109 M ¼ 2:3 nM
437:58 g=mol
Fundamentals of Pharmaceutical Analysis
39
Figure 4.2 Photo of an analytical balance
A normal weighing procedure includes the following steps:
.
Place an empty weighing vessel on the weighing pan.
Reset the reading of the balance (tare to 0.0000 or 0.00000 g).
. Fill the substance or sample to be weighed into the weighing vessel.
. Record the mass on the digital display.
. Clean the balance after use.
.
Calibration is done regularly to ensure high accuracy. Analytical balances calibrate
themselves by placing a standard mass on the pan. It is important never to place drug
substances or samples directly on the pan, as this will contaminate the system. Therefore, the
substances or the sample to be weighed should always be placed in a weighing vessel placed
on the pan. First, place an empty vessel on the pan. Use a paper towel or tissue to handle the
40
Introduction to Pharmaceutical Chemical Analysis
% Relative error
0.00
-0.10
-0.20
-0.30
2
4
6
8
10 12
14
Density of object (g/l)
16
18
Figure 4.3 Weighing error (% relative error) due to the difference in density between the object
to be weighed and the standard mass used for calibration. In this case, the analytical balance was
calibrated with a standard mass of density 8.0 g/l
vessel, because fingerprints will change its mass. Let the balance stabilize for a few seconds,
then reset the digital display to show 0.0000(0) g. Make sure that the weighing vessel is
centered on the pan and that the glass doors protecting the pan are closed to protect it from
drafts. Then place the substance or the sample to be weighed in the weighing vessel, make
sure that the weighing vessel is still centered, close the doors, and wait for the balance to
stabilize before the mass is recorded on the digital display. If some substance or sample is
spilled inside the balance, this should be removed immediately.
Like with many other instruments, errors can occur during operation of the analytical
balance. One source of weighing error occurs if the density of what you are weighing is
different from the density of the standard mass used for calibration. This is because every
time you place an item in the balance, it displaces an equivalent amount of air that also has a
weight. If the difference in density is large between the object to be weighed and the standard
mass used for calibration, the weighing error (% relative error) will be relatively large as
shown in Figure 4.3.
To correct for differences in density, the buoyancy equation can be used:
m1 1 ddair
cw
m2 ¼ ð4:9Þ
1 ddair
where m1 is the measured weight, m2 is the corrected weight, dair is the density of air, dcw is
the density of standard mass used for calibration (typically 8.0 g/ml), and d is the density of
the object considered. Equation (4.9) is not frequently used in the pharmaceutical
laboratory, but it is important to be aware of the principle.
Fundamentals of Pharmaceutical Analysis
41
In addition to this effect it is essential that the object to be weighed has the same
temperature as the standard mass used for calibration. Since calibration normally happens
at room temperature, it is important that the objects you weigh are also kept at
room temperature. A heated object must, therefore, be cooled to room temperature
before it can be weighed. If the object is warmer than room temperature, the weight is
recorded as being too low. Static electricity from the weighing vessel or the object itself
can also be a problem for exact weighing, since the static electricity may affect the
analytical balance.
4.3.2
Pipettes
Pipettes are also standard equipment in the pharmaceutical laboratory. Pipettes are used to
deliver known volumes of liquid. Figure (4.4) shows two different types of pipettes used,
namely the transfer pipette and measuring pipette.
Transfer pipettes are the most accurate. A transfer pipette can only be used to deliver
its specified volume, such as 5 or 10 ml. Transfer pipettes with different volumes are
given in Table 4.1, along with their corresponding tolerance if they belong to class A.
Class A pipettes are the most accurate. If lower accuracy is acceptable, measuring
pipettes can be used. These are graded and can be used to deliver variable volumes, such
as 8.4 ml.
Correct use of a pipette is essential to obtain the highest accuracy. Place the tip of the
pipette into the solution. Use a rubber balloon to suck up the liquid into the pipette. Never use
your mouth to suck up the liquid. The pipette is filled to a level higher than the calibration
mark of the desired volume. Then remove the pipette from the solution and remove traces of
solution on the outer surface of the pipette with paper towel. Hold the pipette tip against the
wall of an empty beaker (waste) and drain the liquid until the meniscus just reaches the
calibration mark. Touching the beaker draws liquid from the pipette without leaving part of a
drop hanging when the liquid reaches the calibration mark. The top of the liquid in the
pipette will form a meniscus as shown in Figure 4.5, and the bottom of the meniscus should
be exactly at the calibration mark of the desired volume.
When the liquid level is adjusted to the mark of the desired volume, the pipette is
transferred to the receiving vessel. The liquid is drained from the pipette by placing the
pipette tip at the glass wall, and allow the liquid to flow slowly along the wall. When
seemingly all the liquid has drained from the pipette, it is still held for a few seconds against
the glass wall before being removed. In the pipette tip, there will still remain a drop of liquid,
but this should not be blown out of the pipette. The liquid remaining in the tip is corrected for
during calibration of the pipette. After using the pipette, it is important to rinse it well before
using it again.
Another type of pipette used in pharmaceutical analysis is the micropipette. A micropipette is shown in Figure 4.6.
Micropipettes are typically used to deliver liquid volumes of between 1 and 1000 ml. By
means of a screw at the end of the micropipette, the volume to be delivered can easily be
adjusted. The tolerance using micropipettes is poorer than when using transfer pipettes and
is typically in the range 1–2%.
To use a micropipette place a fresh tip tightly in the barrel. The tips are made of plastic and
are intended for single use only. Then set the required volume with the screw at the top of the
42
Introduction to Pharmaceutical Chemical Analysis
Figure 4.4 Photo of a transfer pipette and a measuring pipette
Fundamentals of Pharmaceutical Analysis
43
Table 4.1 Tolerance of Class A transfer pipettes
Volume (ml)
0.5
2.0
4.0
10
20
50
Tolerance (ml)
Volume (ml)
Tolerance (ml)
0.006
0.006
0.01
0.02
0.03
0.05
1.0
3.0
5.0
15
25
100
0.006
0.01
0.01
0.03
0.03
0.08
Liquid meniscus
Correct position of the miniscus
at the calibration mark
Calibration
mark
Figure 4.5 Correct position of the meniscus at the calibration mark
pipette. Depress the plunger to the first stop, which corresponds to the desired volume. Hold
the pipette vertically, and dip the tip 3–5 mm into the solution. It is important that the pipette
is held vertically, because the amount that is drawn up in a micropipette depends on the angle
between the pipette and the liquid. Then release the plunger slowly and carefully so that the
pipette tip is filled with the desired volume of liquid. When the pipette tip is to be removed
from the liquid, the tip should slide along the wall to remove excess liquid. This ensures that
liquid located on the outer surface of the pipette tip is removed. Then transfer the liquid
volume to the desired container by placing the pipette tip along the wall, and then push the
plunger gently down to the first stop. After a second pause press the plunger to the bottom
and the remaining liquid is pressed out of the pipette tip. Unlike transfer pipettes and
measuring pipettes, micropipettes must be emptied completely.
All types of pipettes as well as burettes and volumetric flasks, have to be calibrated
regularly to ensure high accuracy. For pipettes, calibration is carried out by pipetting water,
followed by exact weighing of the water with an analytical balance. Using a correction
factor, which takes into account the density of water (which is temperature-dependent) and
Figure 4.6 Photo of a micropipette
44
Introduction to Pharmaceutical Chemical Analysis
Table 4.2 Correction factor for the calibration of pipettes using water
Temperature ( C)
Correction factor (ml/g)
Temperature ( C)
Correction factor (ml/g)
1.0020
1.0023
1.0027
1.0031
1.0035
1.0040
1.0046
1.0051
16
18
20
22
24
26
28
30
1.0021
1.0025
1.0029
1.0033
1.0038
1.0043
1.0048
1.0054
15
17
19
21
23
25
27
29
the buoyancy, the exact volume given by a pipette can be calculated using the following
equation:
Accurate volume ¼ Weight of water Correction factor
ð4:10Þ
The correction factor at different temperatures is given in Table 4.2.
Box 4.10 shows an example of how to calibrate a pipette.
4.3.3
Volumetric Flasks
Volumetric flasks are used for the preparation of solutions with exact concentration. A
volumetric flask is shown in Figure 4.7. A volumetric flask has a narrow neck, and in this
narrow neck is a mark that shows to what exact level the liquid should be filled to get the
exact volume. Volumetric flasks are thus made for a particular volume, and the volume is
given at a particular temperature. This temperature is printed on the flask, and typically it is
20 C.
During preparation of solutions with exact concentration, the first step includes transfer of
an exact amount of the substance into the volumetric flask. This can be either solid material
weighed on an analytical balance, or a liquid delivered with a transfer pipette. It is important
that the substance is transferred to the volumetric flask without losses. Then add some of the
Box 4.10 Example of pipette calibration
A 25-ml transfer pipette is calibrated. Water is pipetted and placed in a weighing vessel.
The mass is determined to 35.2250 g, while weighing the vessel in the dry state displayed
10.2830 g. The temperature in the laboratory was 23 C. At this temperature, the
correction factor was 1.0035 ml/g according to Table 4.2. The exact volume to be
taken with the transfer pipette is calculated as follows:
Weight of water ¼ 35:2250 g 10:2830 g ¼ 24:9420 g
Accurate volume ¼ 24:9420 g 1:0035 ml=g ¼ 25:029 ml
This means that the exact volume of the pipette was 25.029 ml.
Fundamentals of Pharmaceutical Analysis
45
Figure 4.7 Photo of a volumetric flask
solvent and make sure that the substance is dissolved. Finally, add solvent to mark. The
lower meniscus should be level with the mark. Also, volumetric flasks are available in
several different qualities. The quality that gives the best accuracy is often called Class A,
and Table 4.3 shows the tolerance of these with different volumes.
Table 4.3 Tolerances of Class A volumetric flasks
Volume (ml)
1
5
25
100
250
1000
Accuracy (ml)
Volume (ml)
Accuracy (ml)
0.02
0.02
0.03
0.08
0.12
0.30
2
10
50
200
500
2000
0.02
0.02
0.05
0.10
0.20
0.50
46
Introduction to Pharmaceutical Chemical Analysis
Figure 4.8 Photo of a burette
Fundamentals of Pharmaceutical Analysis
47
Table 4.4 Tolerances of Class A burettes
Burette volume (ml)
5
10
25
50
100
4.3.4
Smallest gradation (ml)
Tolerance (ml)
0.01
0.05 or 0.02
0.1
0.1
0.2
0.01
0.02
0.03
0.05
0.10
Burettes
A burette is a long tube of glass with a tap at the lower end. Along the tube there is a
graduation (milliliter scale) that makes it possible to continuously read the volume of liquid
delivered from the burette. Burettes are used in titration (which is discussed in a later
chapter) where a solution is added gradually to a sample until a given point where the
titration is terminated. Then one can read off the consumption of solution with high accuracy
on the burette. A burette is shown in Figure 4.8.
Readings on the burette are made the same way as for pipettes and volumetric flasks. This
means that it is the level of the lower meniscus that is used. When reading the liquid level in a
burette, your eye should be at the same height as the top of the liquid. Also for burettes, there
are several different qualities, and the most accurate are graded as Class A. Tolerances of
Class A burettes are summarized in Table 4.4.
First, the burette is filled with the titration solution. The level of this solution is read on the
milliliter scale before starting. The starting level can be 0.00 ml, or another exact reading
like 0.67 ml. Then the tap is opened gently, and the solution from the burette flows slowly
into the titration vessel. Towards the end of the titration, the titration solution is delivered at
reduced speed by gently opening and closing the tap. At the end of the titration, the level of
the titration solution is read again, and the difference between the start and end level
represents the exact volume consumed during the titration. If the titration was started at
0.67 ml and ended at 15.75 ml, the total volume of solution used during titration was
15.75 0.67 ¼ 15.08 ml. If this titration was performed with a 25-ml burette, with a
tolerance of 0.03, the true value may be between 15.05 and 15.11 ml. In some cases, an air
bubble may be located in the tap before starting, and it is important that this is removed
before the titration begins. This can be done by allowing some titration solution to drain out
of the burette with the tap in the fully open position.
4.4 How to Make Solutions and Dilutions
If you are going to make a solution of a solid with an accurate concentration, you weigh out
the correct mass of the solid on an analytical balance. This is done in a weighing vessel,
where the display of the analytical balance is reset with the empty weighing vessel. The solid
is then carefully placed in the weighing vessel, and the exact mass is displayed on the
analytical balance, normally with four or five digits. Then the substance is transferred from
the weighing vessel and into a volumetric flask. It is important that all the substance is
48
Introduction to Pharmaceutical Chemical Analysis
transferred to the volumetric flask and that no material is lost outside the flask. To remove all
the substance from the weighing vessel, it is normal to wash this with the solvent to be used.
Washing is done so that the flushing solvent flows directly into the volumetric flask. Then
add more solvent to the volumetric flask, but not completely up to the mark. Make sure that
all the substance is dissolved. When all the substance has dissolved, the volumetric flask is
filled up to the mark with solvent. In order not to overfill the volumetric flask, it is common
towards the end to add the solvent from a Pasteur pipette, so that the solvent is filled exactly
to the mark. The exact concentration of the solution can then be calculated. Examples of this
are shown in Boxes 4.1, 4.5, 4.6, and 4.7.
If it is undesirable to wash the weighing vessel to transfer 100% of the solute into the
volumetric flask, the weighing vessel can be weighted after transfer of most of the solute into
the volumetric flask, and the difference in mass from the original recording is the exact
amount of solute transferred into the volumetric flask.
For substances that are liquids at room temperature, one can use the same procedure as for
solids. This means that you weigh an appropriate amount of liquid on the analytical balance.
The liquid is transferred to a volumetric flask with no loss, which is then filled to the mark
with a suitable solvent. In such cases, the concentration of the solution is calculated as
weight per volume, such as mg/ml or %(w/v). Alternatively, liquid substances can be
delivered by a transfer pipette. The liquid is delivered directly from the pipette into a
volumetric flask, and solvent is added to the mark. In such cases, the concentration of the
solution is calculated as volume per volume, such as %(v/v). In all cases when handling
liquids, one must be aware of evaporation. If working with volatile liquids, it is important
always to cap the volumetric flask as soon as possible.
Frequently, solutions with very low concentrations are used in pharmaceutical analysis.
In such cases, you cannot create the solution directly by weighing the substance followed by
a single dilution in a volumetric flask. This is because either very small amounts of material
must weighed out, or it must be diluted in very large volumetric flasks to get the appropriate
concentration. In the former case, the uncertainty is too great, and the solution that is made
will not be accurate in terms of concentration. In practice, it is recommended not to weigh
less than 50–100 mg of substance for the preparation of solutions, as the relative uncertainty
associated with the analytical balance increases with decreasing mass. Similarly, it is
unfavorable to prepare weak solutions by dilution with large solvent volumes. This is mainly
due to the large consumption of solvent.
To prepare very dilute solutions it is common to first make a concentrated solution, and
then to dilute this to a lower concentration. Calculations related to dilutions can be
accomplished with the following dilution equation:
Cundilute Vundilute ¼ Cdilute Vdilute
ð4:11Þ
Cundilute is the concentration of substance in the undiluted solution, Vundilute is the volume of
the undiluted solution, Cdilute is the concentration of substance in the diluted solution, and
Vdilute is the volume of the diluted solution. Cundilute Vundilute represents the number of moles
of substance taken from the concentrated (undiluted) solution, and this equals the amount
placed in the dilute solution (Cdilute Vdilute ). Examples on how to use this equation are shown
in Boxes 4.11 and 4.12.
Fundamentals of Pharmaceutical Analysis
49
Box 4.11 Dilution of hydrochloric acid
A solution of hydrochloric acid at 1.05 M is to be diluted to obtain a solution that is
1.00 102 M. A volume of 500 ml is requested for the diluted solution. How many
milliliters of concentrated solution should be added to a 500-ml volumetric flask?
The dilution equation is used where Cundilute ¼ 1.05 M, Cdilute ¼ 1.00 102 M,
Vdilute ¼ 500.0 ml, and where Vundilute is unknown:
1:05 M Vundilute ¼ 1:00 102 M 500:0 ml
Vundilute ¼
1:00 102 M 500:0 ml
¼ 4:76 ml
1:05 M
Thus, 4.76 ml of 1.05 M HCl should be diluted to 500.0 ml to give a final concentration of
1.00 102 M.
Box 4.12 Dilution of paracetamol solution
50.0 mg paracetamol is dissolved in 100.0 ml methanol in a volumetric flask. From this
solution, 1.00 ml is picked up by a transfer pipette and diluted to 100.0 ml with methanol
in a new volumetric flask. What is the concentration of paracetamol in the final solution?
The first solution of paracetamol will have the following concentration (Cundilute):
Cundilute ¼
50:0 mg
¼ 0:500 mg=ml
100:0 ml
The concentration in the final solution (Cdilute) is calculated according to the dilution
equation, where:
Cundilute ¼ 0:500 mg=ml; Vdilute ¼ 100:0 ml; Vundilute ¼ 1:00 ml :
0:500 mg=ml 1:00 ml ¼ Cdilute 100:0 ml
Cdilute ¼
0:500 mg=ml 1:00 ml
¼ 0:00500 mg=ml ¼ 5:00 103 mg=ml
100:0 ml
4.5 Calibration of Analytical Methods
In quantitative pharmaceutical analysis, it is normal to measure some type of signal (y) in the
analytical instrument, which depends in a well known and reproducible way on the
concentration of analyte in the sample (x). As far as possible, one wishes that y should
be proportional to x, which gives a straight line relationship between y and x, called the
calibration curve:
y ¼ mxþb
ð4:12Þ
50
Introduction to Pharmaceutical Chemical Analysis
Box 4.13 Calculation of exact concentration of CRS in a standard
solution based on reported purity
A standard solution is prepared from omeprazole CRS of reported purity 99.9%.
0.1213 g omeprazole CRS is dissolved in methanol and the volume is adjusted to
100.0 ml. 10.00 ml of this solution is diluted to 100.0 ml with methanol, and 10.00 ml of
this solution is diluted further to 100.0 ml with methanol. The concentration of
omeprazole in the final standard solution is calculated based on the reported purity
and taking the dilutions into consideration (10.00 ml to 100.0 ml and thereafter 10.00 ml
to new 100.0 ml):
ð121:3 mg=100:0 mlÞ ð99:9%=100:0%Þ ð10:00 ml=100:0 mlÞ
ð10:00 ml=100:0 mlÞ ¼ 0:0121 mg=ml
where m is the slope and b is the y-intercept. For all quantitative analysis, it is necessary to
establish the relationship between y and x, and this is called calibration. For unknown
samples, y can be measured and the concentration of analyte can be determined according to
the calibration curve. To calibrate an analytical method it is necessary to purchase highly
pure substance of the analyte. A substance of an analyte with high purity is often termed a
chemical reference substance (CRS). Calibration of an analytical method is performed with
solutions with known concentration of the chemical reference substance. These are
analyzed in the same way as for unknown samples, and the results are plotted as the
calibration curve. The solutions of the chemical reference substance, used for calibration,
are called standard solutions. To ensure an accurate calibration it is important that the
standard solutions are made with great accuracy by the principles discussed in Sections 4.3
and 4.4.
A CRS quality of an analyte is either practically 100.0% pure, or slightly less pure but
with the exact purity known. In the latter case, the exact purity is taken into consideration
when calculating the concentration of the standard solution used for the calibration curve.
An example of this is illustrated in Box 4.13.
4.6 Errors, Accuracy, and Precision
4.6.1
Systematic and Random Errors
In all quantitative pharmaceutical analysis, there will be some uncertainty associated with
the measurements because minor errors can occur in different parts of the procedure. This
means that the analytical result is an estimate for the content (true value) of analyte in a
given sample. It is of course important to reduce uncertainties to a minimum, to ensure that
the analytical results are as close to the true value as possible. Thus, care has to be taken both
during the development of analytical methods and when the analytical methods are used for
quantitative purposes.
Errors made during analytical procedures can be either systematic errors or random
errors. Systematic errors (or determinate errors) are errors that can be detected and which
Fundamentals of Pharmaceutical Analysis
51
therefore can be corrected. Systematic errors will affect all the results in an analytical
series in the same direction, which means that all the answers are either too low or too
high. An example of a systematic error is if a transfer pipette, which is intended to
deliver 10.00 ml, is taking up 10.06 ml. If the pipette is not calibrated, you will not be
aware of this, and the pipette will give a greater volume than expected. This will give rise
to a systematic error and that error will affect any analysis in which the pipette has
been used.
Random errors (or indeterminate errors) are errors that cannot be detected and which
therefore cannot be correct. Random errors affect the results in a random fashion. In some
cases, random errors cause the analytical results to be too high and in other cases, random
errors make the results too low. An example of a random error is by filling pipettes. In
some cases, the meniscus hits the mark exactly, providing an exact volume of liquid.
In other cases, the meniscus is slightly below the mark, resulting in a too low volume, and
in yet other cases the pipette is filled a little more than marked. The error will turn out
different in the different cases, and you are not able to correct the results.
4.6.2
Accuracy and Precision
Systematic and random errors affect analytical measurements, and the analytical results are
basically estimates of the real content (true value). To describe the quality of the estimates
(analytical measurements), the concepts accuracy and precision are used. In everyday life it
is common to use these terms interchangeably, but in pharmaceutical analysis, they have two
completely different meanings. The accuracy expresses how well the analytical result
matches the “true value”. Accuracy can be expressed by the difference between the true
value ðXt Þ and the analytical result ðXÞ, which is the error. The lower is the error or difference
the better is the accuracy. Errors can be given in absolute values or relative values using the
following equations:
Absolute error ¼ X Xt
Relative error ¼
X Xt
100%
Xt
ð4:13Þ
ð4:14Þ
Small errors in an analytical method mean that the method provides good accuracy, the
analytical result is close to the true value, and this is of course preferable.
To establish the accuracy of an analytical method or an analytical measurement, the true
value has to be known, or perhaps it should be called the “true” value. But how can the
“true” value be obtained for unknown samples? There are several ways to do this. One is to
let an experienced analyst do the analysis on the same samples with a well tested
procedure. This may give an indication of the “true” value. Another is to analyze the
same sample with other methods and to compare the results. Good agreement between
several methods gives a good indication, but never a proof that the results are accurate. A
final possibility is to add known amounts of analyte (as a chemical reference substance) to
the sample (standard addition). The analytical results before and after standard addition
can be compared, and this difference can be compared to the exact amount added to
the sample.
52
Introduction to Pharmaceutical Chemical Analysis
High precision
Poor accuracy
High precision
High accuracy
Poor precision
High accuracy
Poor precision
Poor accuracy
Figure 4.9 Illustration of accuracy and precision
Quantitative measurements are always performed by multiple analysis of the same
sample to ensure that results are not affected by random errors. The precision of a method
indicates how close the different measurements are. Precision is expressed by the standard
deviation or relative standard deviation, as defined in Section 4.7.1. High precision means
that the standard deviation is small, which of course is desirable in pharmaceutical analysis.
High precision means that a relatively low number of measurements can be performed on the
same sample without affecting the accuracy of the results significantly. The terms accuracy
and precision are further illustrated in Figure 4.9.
4.7 Statistics
4.7.1
Mean Value and Standard Deviation
Most measurements made in a pharmaceutical analysis follow a Gaussian distribution.
These measurements can be characterized by a mean valueðxÞ and a standard deviationðsÞ.
These terms are defined by the following equations:
x¼
x1 þ x2 þ x3 þ xn
n
ð4:15Þ
Fundamentals of Pharmaceutical Analysis
s¼
vffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
ui¼n
uP
u ðxi x Þ2
ti¼1
n1
53
ð4:16Þ
where x1 ; x2 ; x3 ; . . . xn are the individual measurements and n is the number of measurements. The quantity n 1 is called the degree of freedom. Relative standard deviation (RSD)
is often used instead of standard deviation and is calculated by the following equation:
RSD ¼
s
100%
x
ð4:17Þ
To report an analytical result, which shall always be based on multiple measurements on
the same sample, both the mean value (average) and the standard deviation (or relative
standard deviation) should be given. This is illustrated in Box 4.14.
Box 4.14 Calculation of x, s, and RSD
A quantitative determination of paracetamol in an oral solution, with a specified content
of 24 mg/ml of paracetamol, gave the following results when six individual measurements were conducted:
Measurement 1 ¼ 24.3 mg/ml
Measurement 2 ¼ 23.7 mg/ml
Measurement 3 ¼ 24.7 mg/ml
Measurement 4 ¼ 23.2 mg/ml
Measurement 5 ¼ 23.9 mg/ml
Measurement 6 ¼ 24.2 mg/ml
The mean value is calculated as follows:
x¼
ð24:3 þ 23:7 þ 24:7 þ 23:2 þ 23:9 þ 24:2Þ mg=ml
¼ 24:0 mg=ml
6
The standard deviation is calculated as follows:
s
vffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
u
u ð24:324:0Þ2 þ ð23:724:0Þ2 þ ð24:724:0Þ2 þ ð23:224:0Þ2 þ ð23:924:0Þ2 þ ð24:224:0Þ2
¼ t
mg=ml
61
¼ 0:5 mg=ml
The relative standard deviation (RSD) is calculated as follows:
RSD ¼
0:5 mg=ml
100% ¼ 2:1%
24:0 mg=ml
In the final report, both the mean value of 24.0 mg/ml and the standard deviation of
0.5 mg/ml (or relative standard deviation of 2.1%) should be given to describe the
analytical result.
54
Introduction to Pharmaceutical Chemical Analysis
Table 4.5 Values of Student’s t
Degree of freedom
(n 1)
1
2
3
4
5
6
7
8
9
10
15
20
25
30
40
60
120
1
4.7.2
Confidence level
90%
95%
99%
99.9%
6.314
2.920
2.353
2.132
2.015
1.943
1.895
1.860
1.833
1.812
1.753
1.725
1.708
1.697
1.684
1.671
1.658
1.645
12.706
4.303
3.182
2.776
2.571
2.447
2.365
2.306
2.262
2.228
2.131
2.086
2.068
2.042
2.021
2.000
1.980
1.960
63.657
9.925
5.841
4.604
4.032
3.707
3.500
3.355
3.250
3.169
2.947
2.845
2.787
2.750
2.704
2.660
2.617
2.576
636.619
31.598
12.924
8.610
6.869
5.959
5.408
5.041
4.781
4.587
4.073
3.850
3.725
3.646
3.551
3.460
3.373
3.291
Confidence Intervals
If you make an infinite number of measurements of a sample, and if the measurements are
not affected by systematic errors, the average calculated ðx Þ will be equal to the true value
ðmÞ of the sample, and in a similar way, the standard deviation ðsÞ will be equal to the true
standard deviation ðsÞ. Performing a large number of measurements on a single sample is
not possible, and in practice one is limited to perform only a small number (typically 3–6) of
individual measurements on the same sample. Because of this, it is relevant to calculate the
confidence interval, which is an interval with a given probability that the true value ðmÞ lies
within (assuming no systematic errors). Confidence intervals are calculated using the
following equation:
ts
m ¼ x pffiffiffi
n
ð4:18Þ
where x is the mean value, s is the estimated standard deviation, n is the number of
measurements, and t is taken from Table 4.5. Note that the degrees of freedom are equal to
n 1. A confidence interval calculated at 95% confidence level means that there is 95%
probability that the true value is within the calculated confidence interval.
An example on how to calculate a confidence interval is given in Box 4.15.
Fundamentals of Pharmaceutical Analysis
55
Box 4.15 Calculation of confidence interval
In Box 4.14 a mean value for the content of paracetamol of 24.0 mg/ml and a value of
0.5 mg/ml for the standard deviation were obtained. The number of measurements was six,
and the degrees of freedom were five. A confidence interval for the true value at 95%
confidence level is calculated as follows, utilizing a value of 2.571 for t as found in Table 4.5:
2:571 0:5
pffiffiffi
m ¼ 24:0 mg=ml ¼ ½24:0 0:5 mg=ml
6
This means that, with 95% probability, the true value of the content of paracetamol is
within the interval [24.0 0.5] mg/ml
4.7.3
Comparison of Means with a t-Test
Often you will need to compare two samples and determine whether or not they are
different with respect to the content of an analyte. If the content of paracetamol in one
sample is determined to 24.02 mg/ml and in another sample to 24.36 mg/ml, it is tempting
to believe that the contents of the two samples are different. This may or may not be true
and is entirely dependent on the standard deviation of the two measurements. If the
standard deviations are very small, and thus the method has very good precision, the
content of the analyte is probably different. If, however, the measurements are associated
with larger standard deviations, the measurements of 24.02 and 24.36 mg/ml, respectively, are not different. In order to determine whether two sets of measurements are
significantly different you have to perform a t-test.
A t-test is performed by calculating the mean value for both measurement series, which is
referred to, respectively, as x 1 and x 2 . The number of measurements in series 1 is n1 , and n2 is
the corresponding number of measurements in series 2. First, the pooled standard deviation
(Spooled) for the two measurement series is calculated with the following equation, where s1
and s2 are the standard deviations for the two series of measurements:
sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
s21 ðn1 1Þ þ s22 ðn2 1Þ
spooled ¼
ð4:19Þ
n1 þ n2 2
Then, a tcalculated value is calculated from the following formula:
rffiffiffiffiffiffiffiffiffiffiffiffiffiffi
jx 1 x 2 j n1 n2
tcalculated ¼
spooled
n1 þ n2
ð4:20Þ
The value for tcalculated is then compared with a corresponding t value from Table 4.5, for
example at the 95% confidence level. The number of degrees of freedom is n1 þ n2 2. If
tcalculated is higher than the value found in Table 4.5, it is with 95% probability that the true
values for the two measurement series are different. Similarly, the results are not significantly different if tcalculated is less than or equal to the value found in Table 4.5. An example of
the use of the t-test is shown in Box 4.16.
56
Introduction to Pharmaceutical Chemical Analysis
Box 4.16 Example of t-test
Two oral solutions containing paracetamol have been analyzed, and one wishes to assess
whether the analytical results are significantly different or not at the 95% confidence
level. Oral solution 1 was analyzed with six measurements, where the mean value and the
standard deviation were calculated to 24.02 mg/ml and 1.19 mg/ml, respectively.
Corresponding data for oral solution 2 were 24.36 mg/ml and 0.89 mg/ml, for the
mean value and standard deviation, respectively. First, the pooled standard deviation is
calculated to:
sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
1:192 ð6 1Þ þ 0:892 ð6 1Þ
spooled ¼
¼ 1:05
6þ62
Then, a tcalculated value is calculated to:
j24:02 24:36j
tcalculated ¼
1:05
rffiffiffiffiffiffiffiffiffiffiffi
66
¼ 0:560
6þ6
The number of degrees of freedom are calculated to:
n ¼ n1 þ n2 2 ¼ 6 þ 6 2 ¼ 10
From Table 4.5, the t-value is found to be 2.228. Since tcalculated is below the value in the
table, the analytical results obtained in the two series are not different (95% probability).
4.7.4
Q-Test to Reject Outliers
In some cases, a single measurement in an analytical series deviates significantly from the
rest of the results. Often it is tempting to remove such outliers from the data material. This
can be done if one are sure that a serious error occurred in the experimental work for this
particular and single measurement. If this is not the case, namely that one has an outlier
which cannot be explained, a Q-test must be performed before rejecting the outlier.
To perform a Q-test, the individual measurements are arranged by ascending numerical
value:
x1 ; x2 ; x3 ; . . . xi ; xcritical
ð4:21Þ
where xi is the second largest value and xcritical is the largest value which is requested to be
rejected. Alternatively, if the lowest value is desired to be rejected, the readings are arranged
as follows:
ð4:22Þ
xcritical ; xi ; . . . x3 ; x2 ; x1
where xcritical is the smallest value which is desired to be discarded and xi is the second
smallest value. Then, a Q-value is calculated (Qcalculated ) according to the following
equation:
Qcalculated ¼
jxi xcritical j
jx1 xcritical j
ð4:23Þ
Fundamentals of Pharmaceutical Analysis
57
Table 4.6 Q-values for Q-test
Number of
measurements
Confidence level
3
4
5
6
7
8
9
10
90%
96%
99%
0.94
0.76
0.64
0.56
0.51
0.47
0.44
0.41
0.98
0.85
0.73
0.64
0.59
0.54
0.51
0.48
0.99
0.93
0.82
0.74
0.68
0.63
0.60
0.57
From Table 4.6, read a Q-value based on the number of measurements in the series and the
desired confidence level. If Qcalculated is larger than the table value, the outlier can be
rejected. If Qcalculated is less than or equal to the value found in the table, the outlier cannot
be rejected.
One example of the use of a Q-test is shown in Box 4.17.
4.7.5
Linear Regression with the Method of Least Squares
During calibration, it is common to analyze several standard solutions with different and
well known concentrations of the analyte, and preferably standard solutions are used whose
Box 4.17 Example of Q-test
A quantitative determination of paracetamol in an oral solution, with a specified content
of 24 mg/ml of paracetamol, gave the following results when six individual measurements were conducted:
Measurement 1 ¼ 21.6 mg/ml
Measurement 2 ¼ 23.1 mg/ml
Measurement 3 ¼ 23.2 mg/ml
Measurement 4 ¼ 23.3 mg/ml
Measurement 5 ¼ 23.6 mg/ml
Measurement 6 ¼ 23.7 mg/ml
The value of 21.6 mg/ml is apparently different from the other values. Before this may be
rejected, a Q-test must be performed. The Qcalculated value is calculated as follows:
Qcalculated ¼
j23:1 21:6j
¼ 0:71
j23:7 21:6j
From Table 4.6, the Q-value at 96% confidence level (six measurements) was found to be
0.64. Since Qcalculated is greater than the Q-value from the table, the outlier can be
rejected.
58
Introduction to Pharmaceutical Chemical Analysis
concentrations cover the range within which the real samples are expected. Then the
measured signal for the different standard solutions (from the analytical instrument) is
plotted as a function of analyte concentration, and as mentioned earlier, there will normally
and preferably be a straight-line relationship between concentration and signal. A straight
line is described using the following equation:
y ¼ mxþb
ð4:24Þ
where m is the slope and b is the y-intercept. If you have n data points, representing n
different standard solutions with different concentrations, that generally can be written in
the form ðxi ; yi Þ, one can calculate the slope m and y-intercept b based on the least squares
method according to the following equations:
P
P P
n ðxi yi Þ xi yi
ð4:25Þ
m¼
D
P
P
P 2 P
ðxi Þ yi ðxi yi Þ xi
b¼
ð4:26Þ
D
D is calculated from the following equation:
X 2
X
D¼n
ðxi Þ2 xi
ð4:27Þ
Thus, based on our n data points obtained for the standard solutions, the best fit for a linear
relationship can be established between measured signal and analyte concentration by using
Equations (4.24), (4.25), (4.26), and (4.27). To express how well our n data points for the
standard solutions fit the linear relationship, the square of the correlation coefficient (R2) is
calculated by the following equation:
P
½ ðxi x Þðyi y Þ2
2
R ¼P
ð4:28Þ
P
ðxi x Þ2 ðyi y Þ2
where x is the mean of all the x values and y is the mean of all the y values. R2 should be very
close to 1.0 to represent a linear fit, and normally values above 0.995 are considered as
acceptable for linear correlation. Linear regression using the least squares method and
calculation of the square of the correlation coefficient are normally and very easily
performed on computers using Excel software. Box 4.18 shows an example of calculating
a linear calibration curve based on the least squares method.
4.7.6
How to Present an Analytical Result
As mentioned earlier, an analytical result should always be based on several measurements.
The analytical result should preferably be reported as a confidence interval, as discussed in
Section 4.7.2. Alternatively, the mean value can be reported together with the standard
deviation associated with the measurement. Any outliers should be removed before
calculating the confidence interval, the mean, and the standard deviation. It is highly
important to present the analytical result with the correct number of digits. In cases where
the precision is high, several digits have to be included in the analytical result, while results
Fundamentals of Pharmaceutical Analysis
59
Box 4.18 Calculation of calibration curve based on the least
squares method
Four standard solutions are analyzed and used to establish a standard curve. The
following values were obtained:
Concentration of standard
(xi; mM)
Signal of standard
(yi; mV)
1.00
2.00
3.00
4.00
2.33
4.84
7.10
9.20
Using Excel, the calibration equation [used to convert the measured signal in mV (y) to
the concentration of analyte in mM (x)] is calculated to:
y ¼ 2:29x þ 0:150
R2 is calculated to 0.9984. Try to use Equations (4.24) to (4.28) to verify these results.
from methods with low precision should be presented with fewer digits. The following setup
can be used to determine the number of digits in an analytical result:
Calculate the mean value ðxÞ and standard deviation (s) for the set of measurements.
Enter ðxÞ in the form a 10n , where a is a number between 1 and 10, and n is an integer.
. Write (s) in the form b 10n , where n has the same value as above.
. Align b under a. Draw a vertical line to the right of the first digit in b different from 0. Then
remove any digits in a to the right of this line.
. The number of reliable digits in the result is the number of digits that are left in a.
. If the first digit removed from a is 5, the last digit in a should be rounded upwards.
.
.
Box 4.19 shows an example of how to adjust the analytical result with the correct number of
digits.
The procedure outlined above is in agreement with the significant figure convention. This
convention states that all numbers should be given only with significant figures, which are all
the digits known with certainty plus the first uncertain digit. With a 50-ml burette as
example, with graduations at every 0.1 ml, it is easy to see that the liquid level is greater than
24.5 ml, but less than 24.6 ml (Figure 4.10). Thus, all the digits of 24.5 ml are certain.
However, the position of the liquid between the graduations can be estimated to perhaps
0.02 ml, and in this particular example it is estimated to 24.56 ml. The last digit is the first
uncertain digit. Thus, the reading should be reported as 24.56 ml, which contains four
significant figures. The reading can also be reported as 0.02456 l, but still it contains four
significant figures after the zeros.
Significant figures are important to understand and this should be discussed in more
detail. The following numbers 1036, 103.6, 10.36, 1.036 all contain four significant figures.
60
Introduction to Pharmaceutical Chemical Analysis
Box 4.19 Adjustment of analytical result to correct number of digits
The following results were obtained in Box 4.17 for a quantitative measurement of an
active pharmaceutical ingredient in tablets (mg per tablet):
23:34
23:56
23:12
23:17 23:67
21:56
As mentioned in Box 4.17, the value of 21.56 was rejected based on a Q-test, and the
following measurements remained:
23:34 23:56
23:12
23:17
23:67
For this series the mean value was calculated to be 23.3720 mg per tablet. The standard
deviation was 0.2395 mg per tablet. Adjustment of the analytical result to the correct
number of digits was accomplished as follows:
x ¼ 2:33j720 101
s ¼ 0:02j395 101
The result should therefore be printed as 23.4 mg per tablet with a standard deviation of
0.2 mg per tablet. Alternatively, the result can be printed as 23.4 0.3 mg per tablet as a
95% confidence interval.
24.6
24.5
Figure 4.10 Reading of a burette
The numbers 1036 and 1.036 103 also contain both four significant figures. The numbers
0.001036, 0.01036, 0.1036, 1.036, 1.036 103, and 10.36 102 also contain four
significant figures. To summarize:
.
.
Zeros before the decimal point are not significant.
Zeros between non-zero digits are significant.
Fundamentals of Pharmaceutical Analysis
61
Terminal zeros may or may not be significant. For example, if the volume of a beaker is
expressed as 1.0 l, the presence of the otherwise unnecessary zero implies that the volume
is known to a few tenths of a liter. Both 1 and 0 are significant figures. If the same volume is
reported as 1000 ml, the last two zeros are not significant because the uncertainty here is still
a few tenths of a liter, corresponding to a few hundred milliliters. In order to follow the
significant figure convention in this case, scientific notation should be used and the volume
reported as 1.0 103 ml.
Care is required in determining the appropriate number of significant figures in values
obtained from arithmetic combinations of two or more numbers. The following rules can be
used:
.
Addition and subtraction: significant figures after the decimal place not higher than those
of that number having the fewest significant figures after the decimal point;
. Multiplication or division: number of significant figures equal to the smallest number
carried by any of the value being multiplied and/or divided;
. Taking logarithms: quote the logarithm with the mantissa having as many figures as the
significant figures in the original number.
Examples of determining the appropriate number of significant figures are shown in
Box 4.20.
Box 4.20 Significant figures
Calculate the following:
5:4 ml þ 0:042 ml þ 8:36 ml
This equates to 5.4 ml þ 0.042 ml þ 8.36 ml ¼ 5.4 ml þ 0.042 ml þ 8.36 ml ¼
13.802 ml ¼ 13.8 ml.
The answer should be 13.8 ml because 5.4 ml determines the number of significant
figures after the decimal place (¼ 1). Note that rounding of the data was accomplished at
the end of the calculation to avoid loss of information. Digits that are not significant
figures can be written as subscript in order to distinguish them from significant figures.
Calculate the following:
24 ml 4:52 M
¼ 1:0848 M ¼ 1:1 M
100:0 ml
The answer should be 1.1 M because 24 ml determines the number of significant figures
(¼ 2).
Calculate the following:
logð4:321 106 Þ ¼ 5:3644157 ¼ 5:3644
The answer should be 5.3644 with four significant figures after the decimal place
because the original number contained four significant figures.
62
Introduction to Pharmaceutical Chemical Analysis
4.8 Some Words and Concepts
Earlier in this chapter, the words analyte, matrix, qualitative analysis, quantitative analysis,
accuracy, precision, calibration, chemical reference substance, and standard solutions
were discussed. In this final section, some additional terms and concepts are discussed
that are important in pharmaceutical analysis.
4.8.1
Analysis and Determination
In pharmaceutical analysis, it is common to use the terms analysis and determination. The
word analysis is used in connection with the samples to be examined, while the word
determination refers to the substance that you want to measure quantitatively. Thus, blood
samples, urine samples, raw materials, and tablets are analyzed, while one determines
nortriptyline in blood, amphetamine in the urine, paracetamol in raw materials, and
diazepam in tablets. It is important that these terms are used correctly.
4.8.2
Sample Replicates and Measuring Replicates
For a given sample, it is common to subject several small portions to analytical measurement, and different portions of the same sample are called sample replicates. The purpose of
analyzing several sample replicates is to ensure that the analytical result is representative for
the entire sample. For each sample replicate, it is normal to perform several individual
measurements, and these measurements are called measurement replicates. The purpose of
taking several measurement replicates is to ensure that the final analytical result is not
affected by serious random errors.
4.8.3
Interference
Quantitative pharmaceutical analysis is normally performed by measurement of a physical
parameter ( y), which in a well known and reproducible way depends on the concentration of
analyte in the sample (x). In some cases, substances other than the analyte may contribute to
y, and this is called interference. Interferences are very important to eliminate, as they give
rise to errors in quantitative measurements. During the development of analytical methods,
it is therefore of crucial importance to ensure that the method is not affected by interference
from other substances in the sample. This is accomplished by analyzing blind samples as
defined below.
4.8.4
Blind Samples
A blind sample is a sample which does not contain the substance or substances to be
determined, but all or as many as possible of the other ingredients in the sample. Blind
samples undergo the same analysis procedure as the samples. Blind samples are used to
control that other substances in the samples do not affect the analytical result. In other
Fundamentals of Pharmaceutical Analysis
63
words, blind samples are used to check that the analytical method does not suffer from
interferences. If you want to analyze drugs in pharmaceutical preparations, it is common to
use a placebo preparation as blind sample, which is a product that contains all the ingredients
except the active pharmaceutical ingredient. For the analysis of drugs in plasma samples
and urine samples, it is common to use, respectively, drug-free plasma and urine as the
blind sample.
5
Titrimetric Methods
This chapter discusses the principles of quantitative analysis based on titrimetric methods.
Titrimetric methods are official methods in the European Pharmacopeia and are used for
quality control of active pharmaceutical ingredients (APIs) and excipients. The chapter
focuses on titrations based on acid–base reactions and redox reactions and also gives an
overview of various principles of endpoint detection. Examples of titrimetric methods for the
quantitative determination of APIs and excipients are discussed in Chapter 21.
5.1 Introduction
In titrimetric methods the volume of a reagent needed to react with an analyte is normally
measured. To the analyte solution are added increments of a reagent solution (titrant) of a
precisely known concentration until the reaction between the reagent and the analyte is
complete. The titrant is added either manually from a burette (see Figure 5.1) or from an
automatic titration apparatus, so that the volume of titrant is controlled at all times during
the titration.
A titrimetric reaction can be illustrated by the following equation:
x ½Analyte þ y ½Titrant ! Products
ð5:1Þ
where x and y is the number of moles of analyte and titrant needed to complete the titration.
Titration is finished when virtually all the analyte has reacted and is transferred to products.
The titration has reached the equivalence point. The amount of reagent consumed is read on
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
66
Introduction to Pharmaceutical Chemical Analysis
Burette
Titrant
Stopcock
Analyte solution
Magnetic stirring bar
Magnetic stirrer
Figure 5.1 Burette used for titration
the burette, and the quantity of analyte in the sample solution can be calculated on the basis
of knowledge of the values of x and y. An example of such a calculation is shown in Box 5.1.
A number of requirements should be fulfilled to make a titration feasible, and these may
be briefly summarized as follows:
1. The titration reaction must be well defined without side reactions.
2. The reaction must be virtually complete (approximately 100% of analyte must be
converted to product).
3. Other substances in the sample must not react with the reagent.
4. The reaction rate should be high.
5. There must be a method to detect when the equivalence point is reached.
6. The exact concentration of reagent must be known.
Point 1 highlights that, in titrimetric reactions, we need to know the exact amount of titrant
that reacts with each mole of analyte. This requirement is necessary to calculate the amount
Titrimetric Methods
67
Box 5.1 Calculation of the concentration of analyte in a titrimetric
method
A sample solution of H2SO4 of unknown concentration was titrated with a solution of
1.056 M NaOH to determine the molar concentration of H2SO4. 25.00 ml of the sample
solution was titrated to the equivalence point with 23.55 ml of NaOH solution. H2SO4
and NaOH react according to the following reaction equation:
H2 SO4 þ 2NaOH ! SO4 2 þ 2Na þ þ2H2 O
The number of moles of NaOH consumed is calculated as follows:
nNaOH ¼ ð23:55 ml 1:056 MÞ=ð1000 ml=lÞ ¼ 2:49 102 mol
Since H2SO4 and NaOH react in a molar ratio of 1 : 2, the number of moles of H2SO4 in
the sample solution (nH2SO4) was 1.24 102 mol. The molar concentration of H2SO4 in
the sample must therefore have been:
CH2SO4 ¼ ð1:24 102 mol 1000 mlÞ=ð25:00 mlÞ ¼ 0:497 M
of analyte in the sample, as exemplified in Box 5.1. This requirement also means that the
analyte and the reagent must react in a well defined manner without the possibility of side
reactions. Criteria 2 and 3 are also very fundamental. If the reaction does not progress to
almost 100% or if other drug substances also consume the titrant, the consumption of
reagent cannot be correlated to the amount of analyte in the sample solution. Criterion 4 is
not essential, but it is an advantage that the reaction between the analyte and the reagent is
fast, so that the titration can be performed in a short time. Criterion 5 states that we must have
a method to detect when the equivalence point is reached, so that we can end the titration and
read the consumption of titrant for calculation of the concentration of analyte. As shown in
Box 5.1, these calculations assume that the concentration of titrant is known (Criterion 6).
Reactions used in titrimetric methods may be of several types, and in this chapter, the
following types are described:
.
Acid–base reactions (acid–base titration);
Reduction–oxidation reactions (redox titration);
. Complexometric reactions (complexometric titration).
.
The emphasis will be placed on acid–base and redox titrations, while complexometric
titrations will only be discussed briefly. In addition to characterizing the titration according
to the chemical principle as the basis for the reaction, it is customary to characterize the
titration by the method used to detect the equivalence point of titration. Of the most
important detection methods are visual detection (color change) and detection using
potentiometry.
During the titration the titrant is added to the solution until the equivalent point is reached,
where virtually all the analyte has been converted into products. We do not write that 100%
of the analyte has been converted into products due the fact that all titrimetric reactions are
equilibrium reactions which means that Equation (5.1) actually should be written:
68
Introduction to Pharmaceutical Chemical Analysis
x ½Analyte þ y ½Reagent $ Products
ð5:2Þ
The reactions selected in titration have equilibriums completely shifted to the right. The
titration is stopped when the endpoint of titration is reached. The endpoint is the point at
which the operator sees that the titration ends. Ideally, the endpoint coincides with the
equivalence point of titration. If no compliance, the result is encumbered with a titration
error. Section 5.2 discusses how to choose a reasonable endpoint detection to avoid errors.
To achieve high accuracy in titration, it is essential to know the exact concentration of
titrant. This is determined in advance of the titration process. To achieve this it is necessary
to have a primary standard. A primary standard is a reagent with the following characteristics:
.
Has a high chemical purity (>99.95%);
Has a well defined chemical composition (including the amount of crystal water);
. Does not change the composition during storage or in contact with air and light;
. Must have good solubility in the titration solvent (water or organic solvent);
. Should be involatile;
. Should have a relatively high molecular weight;
. Should be a strong electrolyte.
.
A variety of materials meet these requirements, such as HOOCC6H4COOK (potassium
hydrogen phthalate), Na2CO3 and K2Cr2O7. Titrant solutions of primary standards can be
made directly by accurate weighing of the standards and subsequent dissolution and dilution
to a specific volume. Such solutions will have an accurately known concentration. In some
cases, such solutions are used directly to titrate the analyte. In other cases, however, it is not
possible to obtain a primary standard that are suitable for direct titration. In this case the
solution of titrant is first standardized with a solution of a primary standard to determine the
exact concentration of the titrant. Such procedures are called standardization of titrant
solutions. An example is shown in Box 5.2 for HCl.
Box 5.2 Standardization of HCl with sodium carbonate
A solution with a concentration of HCl at about 0.1 M must be standardized to determine
the exact concentration of HCl. Sodium carbonate is used as primary standard. An
amount of 0.1006 g of Na2CO3 (molecular weight ¼ 105.99 g/mol) was dissolved in
water and diluted to 20 ml. This solution was titrated with 0.1 M HCl, and it consumed
18.55 ml of the solution to be standardized. The following reaction occurs:
2HCl þ CO3 2 ! 2Cl þ H2 CO3
The number of moles of primary standard in the solution (nPS) is calculated as follows:
nPS ¼ ð0:1006 gÞ=ð105:99 g=molÞ ¼ 9:492 104 mol
Since the primary standard and HCl react in the molar ratio 1 : 2, the amount of HCl
added was 1.898 103 mol. The molarity of HCl solution (MHCl) is calculated as
follows:
MHCl ¼ ð1:898 103 mol 1000 ml=lÞ=ð18:55 mlÞ ¼ 0:1023 M
Titrimetric Methods
69
Burette
Voltmeter
Indicator
electrode
Reference
electrode
Magnetic stirrer
Figure 5.2 Principle of potentiometric endpoint detection
Primary standards suitable for the various types of titration are discussed later.
The endpoint of a titration may be determined using an indicator, which is added to the
analyte solution. The indicator has a property (usually color) that changes abruptly near the
endpoint. The change is caused by the appearance of excess titrant or disappearance of
analyte. Color indicators are much in use, and their behavior is discussed in the sections on
acid–base titration and redox titration. The advantage of color indicators is that they are low
in price and very easy to use. Potentiometric titration is an alternative to color indicators for
endpoint detection. The advantage of potentiometry is that the technique allows for
automation and sometimes provides more sensitive endpoint detection. The principle of
potentiometric detection is shown in Figure 5.2.
A two-electrodes system is placed in the titration solution, associated with a voltmeter.
One electrode acts as an indicator electrode, and the potential of the indicator electrode
changes as a function of changes in analyte concentration.The other electrode is called the
reference electrode, and the potential of this does not change during titration. The voltmeter
measures the difference between the potential of the indicator electrode and the reference
electrode, and this potential is plotted as a function of the added amount of titrant. The
resulting titration curve is used to determine the endpoint of titration.
Several electrodes are used within potentiometric detection. In this book the principle of
the silver–silver chloride electrode, the calomel electrode, the glass electrode, the combined
glass electrode and the platinum electrode is briefly reviewed. The silver–silver chloride
70
Introduction to Pharmaceutical Chemical Analysis
Ag wire
Saturated solution of
KCl and AgCl
Solid KCl and AgCl
Porous glass
Figure 5.3 Principle of the silver–silver chloride electrode
electrode and the calomel electrode act as a reference electrode. The former is schematically
shown in Figure 5.3.
The electrode consists of a tube filled with a saturated solution of KCl and AgCl, and a
silver wire is placed in the solution. At the bottom of the tube is a plug of porous glass so that
the electrode has electrical contact with the titration solution. The electrode reaction can be
described by the following equation:
AgClðsÞ þ e $ AgðsÞ þ Cl
ð5:3Þ
The standard reduction potential (E 0) at the electrode is 0.199 V.
The calomel electrode is an alternative to the silver–silver chloride electrode. The
calomel electrode is also a reference electrode, where Hg/Hg2Cl2 replaces Ag/AgCl. The
reaction is:
Hg2 Cl2 ðsÞ þ 2e $ 2HgðlÞ þ 2Cl
0
ð5:4Þ
The standard reduction potential (E ) of the electrode is 0.244 V.
Of the indicator electrodes the glass electrode is very important since this is used for the
measurement of pH during titration. A glass electrode and its reference electrode are shown
in Figure 5.4.
The glass electrode consists of a pH-sensitive glass membrane shaped like a bubble at the
bottom of the electrode. Inside this bubble, there is a solution of HCl saturated with silver
Titrimetric Methods
71
Voltmeter
Silver wire
Reference
electrode
Sample solution
Glass membrane
Hydrochloric acid
saturated with AgCl
Figure 5.4 Principle of pH measurements with a glass electrode and a reference electrode
chloride. This solution is in contact with a silver wire connected to a voltmeter, acting as a
silver–silver chloride electrode. A reference electrode, which can be either a silver–silver
chloride electrode or a calomel electrode, is located in the solution next to the glass electrode.
It may seem strange that two reference electrodes can serve as an indicator electrode for Hþ
ions. The reason for this is that the pH-sensitive glass membrane forms negative charges on the
glass and the number of negative charges increases with decreasing Hþ concentration. The
number of negative charges on the inside of the pH-sensitive glass membrane is constant,
while the charge on the outside varies with pH in solution. There is a charge difference across
the membrane, and this charge difference is measured using two silver electrodes connected to
the voltmeter. Often a glass electrode and a reference electrode are combined in one cell,
called a combined glass electrode. However, it is important to emphasize that the combined
glass electrode is still made up of two reference electrodes as mentioned above.
In some types of titration (i.e., redox titration) it is also necessary to use a platinum
electrode together with a reference electrode. Each of these consists of a platinum wire.
Platinum is inert and acts as an electrical conductor. Compared to color indicators
potentiometric endpoint detection has a slightly slower response time. That is, there is
often a slight delay from the addition of titrant until a stable signal on the voltmeter is read.
For practical work, this means rapid titration until approximately 1 ml before the endpoint.
72
Introduction to Pharmaceutical Chemical Analysis
Then the titrant is added slowly, so that equilibrium in the system is achieved between each
addition of titrant and voltmeter reading.
5.2 Acid–Base Titrations
Acid–base titrations are based on acid–base reactions and are used to titrate solutions with
unknown concentrations of either basic or acidic analytes. Strong acids and bases are
entirely dissociated in water, and the titration reaction is therefore:
H3 Oþ þ OH ! 2H2 O
ð5:5Þ
If you measure the pH of the solution when a strong acid is gradually titrated with a strong
base, you will get a titration curve as shown in Figure 5.5.
As can be see from the curve, pH is very low in the sample solution by the start of the
titration (pH 0), and pH rises only slightly as the base is gradually added. Just before
the equivalence point pH begins to change sharply, and if the titration is continued after
the equivalence point, the curve levels out at high pH values. The equivalence point is the
steepest point of the curve. It is the point at maximum slope and is therefore an
inflection point.
pH
14
12
Equivalence point
10
Color change for
phenolphthalein
8
6
Color change for
methyl orange
4
2
ml
20
40
volume of 1 M NaOH
Figure 5.5 Titration curve for titration of 25 ml of 1.0 M HCl with 1.0 M NaOH
Titrimetric Methods
OH
73
OH
OH
O
O
COOH
O
Figure 5.6 Structural formula of phenolphthalein and structural rearrangements responsible for
color change
To detect the endpoint of titration a color indicator is often added to the sample solution to
perform visual endpoint detection. Color indicators of acid–base titration are acids or basic
substances, which change color as the indicator is in its acid or basic form, respectively. An
example of an acid–base indicator is phenolphthalein (PP), which is shown in Figure 5.6.
Phenolphthalein has a pKa value of 9.4. This means that, at pH 9.4, 50% of the indicator is
in the acid form and 50% in the base form, and the indicator is colored red. At pH 8.4 only
10% of the indicator is in the base form, and the red color is very weak, while at pH 10.4 90%
of the indicator is in the base form and the red color is very strong. In practice, the color
change of PP is clearly seen between pH 8.4 and 10.4. We see from the titration curve in
Figure 5.5 that the pH rises very rapidly from about pH 4 to 12 at the equivalence point. If PP
is added to the sample solution the color changes from colorless to red at the equivalence
point. This color shift is the basis to stop the titration and thus determine the endpoint.
Generally, the useful range for color indicators is 1 pH unit on either side of its pKa value.
There are a variety of acid–base indicators that provide a suitable color shift for detecting the
endpoint in an acid–base titration. Table 5.1 shows a few examples.
We see from the titration curve in Figure 5.5 that pH increases rapidly at the equivalence
point, and other acid–base indicators with other pKa values can also be used. An example is
methyl orange (MO), which has a pKa value of 3.5.
In Figure 5.5 we see the titration curve when we titrate a strong acid (HCl) with a strong
base (NaOH), where the pH rises very rapidly to high values at the equivalence point. If we
Table 5.1 Color indicators for acid–base titrations
Indicator
Methyl orange
Bromophenol blue
Methyl red
Chlorophenol blue
Cresol red
Phenolphthalein
pKa
3.5
4.0
5.1
6.0
8.3
9.4
Color change
Color
pH
Acid
Base
2.5–4.7
3.0–5.0
4.1–6.1
5.0–7.0
5.3–9.3
8.4–10.4
Red
Yellow
Red
Yellow
Yellow
Colorless
Yellow
Purple
Yellow
Red
Red
Red
74
Introduction to Pharmaceutical Chemical Analysis
pH
14
Equivalence point
12
Color change for
phenolphthalein
10
8
6
Color change for
methyl orange
4
2
ml
20
40
volume of 1 M NaOH
Figure 5.7 Titration curve for 25 ml of 1.0 M solution of acetylsalicylic acid with 1.0 M NaOH
instead titrate a weaker acid with a strong base, the pH rise at the equivalence point is less
pronounced. This is shown in Figure 5.7, where 25 ml of 1.0 M acetylsalicylic acid is titrated
with 1.0 M NaOH.
In this case, PP can still be used as color indicator since the color shift falls within the
range of the inflection of the titration curve. Methyl orange (MO) cannot be used to detect
the endpoint in this case because its color shift falls outside the inflection point of the
titration curve. In general, pH variations at the equivalence point are less prominent for the
weaker acids and bases, as shown in Figure 5.8. The consequence is that it is difficult to
detect the endpoint of titration by using color indicators. As a rule of thumb, acids with
pKa > 6.0 and bases with pKa < 8.0 cannot be titrated with color indicators for endpoint
detection. These limits may be extended somewhat if one uses potentiometric endpoint
detection, but generally weak acids and bases must be titrated in organic solvents instead of
water. This is discussed in more detail in Section 5.3.
Some acids and bases include groups that can donate or accept more than one proton. This
means that one mole of analyte is equivalent to more than one mole of titrant. When the pKa
values for the various groups differ by more than four units, the titration curve of that
compound has more than one inflection point. Sodium carbonate is an example of a diprotic
base, which can be titrated with hydrochloric acid, according to the following reactions:
CO3 2 þ Hþ ! HCO3
ð5:6Þ
Titrimetric Methods
75
pH
14
12
pKa = 10
10
pKa = 8
8
pKa = 6
6
pKa = 4
4
pKa = 2
Strong acid
2
ml
25
50
volume of 1 M NaOH
Figure 5.8 Titration curve for 50 ml of 0.1 M acid of varying pKa values with 0.1 M NaOH
HCO3 þ Hþ ! H2 CO3
ð5:7Þ
The titration curve for titration of 25 ml of 1 M sodium carbonate with 1 M HCl is shown in
Figure 5.9. Phenolphthalein can be used to detect the first inflection point, while MO can be
used to detect the second inflection point. Similarly, it is possible to determine two analytes
in the sample solution if their pKa values differ by at least four units.
Acidic drugs are titrated with basic titrants, typically with solutions of NaOH. NaOH is
not a primary standard and NaOH solutions must be standardized in advance. According to
the European Pharmacopeia (Ph.Eur.), solutions of NaOH are standardized by titration with
solutions of HCl. HCl is not a primary standard and must be standardized with Na2CO3 as
shown in Box 5.2. Alternatively, solutions ofNaOHmay be standardized by direct titration
with potassium hydrogen phthalate. For titration of basic pharmaceuticals, a solution of HCl
is used as titrant. According to Ph.Eur., solutions of HCl are standardized with sodium
carbonate as shown in Box 5.2.
5.3 Acid–Base Titrations in Non-Aqueous Media
Acid–base titrations are successful in aqueous environments when the acid or base to be
titrated is not too weak. In practice this means that the pKa of acids should not be higher than
76
Introduction to Pharmaceutical Chemical Analysis
pH
14
12
10
Color change for
phenolphthalein
8
6
Color change for
methyl orange
4
Equivalence points
2
ml
20
40
volume of 1 M HCl
Figure 5.9 Titration curve for titration of 25 ml of 1.0 M sodium carbonate with 1.0 M HCl
6.0, and similarly for bases, the pKa should not be higher than 8.0 when using color
indicators for endpoint detection. With potentiometric endpoint detection, acids with pKa
values of up to 8.0 and bases with pKa values down to 6.0 can be titrated in an aqueous
environment. If even weaker acids or bases are titrated in an aqueous environment, water,
which itself can act as both a weak base and a weak acid, could compete for the titrant.
Water is protonated according to the following reaction:
H2 O þ Hþ ! H3 Oþ
ð5:8Þ
and may compete in the titration of a weak base:
RNH2 þ Hþ ! RHþ
3
ð5:9Þ
Water is deprotonated by the following reaction:
H2 O þ B ! OH þ BHþ
ð5:10Þ
and may compete in the titration of a weak acid:
RCOOH þ B ! RCOO þ BHþ
ð5:11Þ
Titration of weak acids and bases must therefore take place in a solvent other than water and
preferably in solvents that enhance their acidic and basic character. For weak bases, which
constitute the main group in connection with drugs, it is customary to carry out the titration
Titrimetric Methods
77
in anhydrous acetic acid instead of water. Anhydrous acetic acid is a very weak proton
acceptor and weak bases are strong proton acceptors. Therefore, anhydrous acetic acid does
not compete with weak bases for protons. Only very strong acids are able to protonate
anhydrous acetic acid according to the following equilibrium:
CH3 COOH þ HA $ CH3 COOH2 þ þ A
ð5:12Þ
Among the most common acids, perchloric acid is the strongest in anhydrous acetic acid
(stronger than, say, HCl), and it is usual to titrate weak bases with perchloric acid dissolved
in anhydrous acetic acid. Weak bases (B) are titrated according to the following reaction:
B þ CH3 COOH2 þ ! BHþ þ CH3 COOH
ð5:13Þ
The substances to be titrated are often dissolved in anhydrous acetic acid, dioxane or other
organic solvents that are soluble in anhydrous acetic acid. It is important that no trace of
water is present in the solution. Addition of acetic anhydride, which hydrolyzes to acetic
acid, removes residual water after the following reaction:
ðCH3 COÞ2 O þ H2 O ! 2CH3 COOH
ð5:14Þ
The endpoint is determined by the addition of a color indicator, and cresol red, methyl red
and crystal violet are often used. Several of these are also used in aqueous titration, but it is
important to emphasize that the color change for the same indicator in water and anhydrous
acetic acid is different since they have different pKa values in water and anhydrous acetic
acid. Indicators for non-aqueous titration must be dissolved in a suitable organic solvent, not
in water.
Potentiometric endpoint detection is preferred to color indicators. In a water-free
environment a glass electrode is used as indicator electrode, while the reference electrode
is a silver wire coated with AgCl. The reason why you cannot make use of silver–silver
chloride, calomel or combined glass electrodes, is that these leach out small amounts of
chloride ions and water, both of which affect the water-free titration.
For weakly basic pharmaceutical compounds in salt form (acting as acidic compounds),
anhydrous titration with perchloric acid in anhydrous acetic acid is used to determine the
amount of counter ion that is in the analyte, since protonated acetic acid reacts with
negatively charged ions. Titration in such cases, indirectly and based on the stoichiometric
balance, also provides an accurate measure of analyte purity. Such titrations of counter ions
work well for counter ions like tartrates, acetates, phosphates and sulfates. When the analyte
is a hydrochloride or hydrobromide, protonated anhydrous acetic acid does not react with
the negatively charged ions. In these cases titration is done with perchloric acid dissolved in
formic acid according to the following reaction:
Cl þ HCOOH2 þ ! HCl þ HCOOH
ð5:15Þ
Unlike aqueous titration, non-aqueous titration is significantly dependent on temperature.
This is because organic solvents have a substantially higher volumetric thermal expansion
coefficient than water. For practical work, this means that it is important to measure the
temperature when the titrant is standardized (t1) and to measure the temperature when the
titration is done (t2). If temperatures are different, you have to do a volume correction for the
volume of titrant (V ), and the following formula applies to the calculation of the proper
78
Introduction to Pharmaceutical Chemical Analysis
Box 5.3 Volume correction by titration with 0.1 M perchloric acid in
anhydrous acetic acid
Standardization of the titrant was at 20 C. The titration was at 25 C and one of the
titrations consumed 15.45 ml of titrant. Since the temperatures differs the volume of
titrant must be corrected:
Vk ¼ 15:45 ml ½1 þ ð20 25Þ 0:0011 ¼ 15:35 ml
Because of the temperature increase, the volume of titrant is changed by:
½ð15:45 15:35Þ=15:35 100% ¼ 0:52%:
This is a significant difference and highlights the importance of performing volume
correction.
titration volume (Vk) if you use 0.1 M perchloric acid dissolved in anhydrous acetic acid:
Vk ¼ V ½1 þ ðt1 t2 Þ 0:0011
ð5:16Þ
Box 5.3 shows an example of volume correction by titration with 0.1 M perchloric acid in
anhydrous acetic acid.
Perchloric acid in anhydrous acetic acid must be standardized before it can be used as a
titrant. According to Ph.Eur. this is done by titrating a known quantity of potassium
hydrogen phthalate according to the following reaction:
HCOOCC6 H4 COO þ CH3 COOHþ
2 ! HOOCC6 H4 COOH þ CH3 COOH
ð5:17Þ
5.4 Redox Titrations
Acid–base titrations are based on the transfer of protons. In addition to this type of titration it
is common also to take advantage of reactions which transfer electrons. Such reactions are
called redox reactions and, when used in titration, this is called a redox titration. A redox
titration can be illustrated by the following equation:
Ox1 þ Red2 ! Red1 þ Ox2
ð5:18Þ
Where the subscript 1 refers to the substance to be titrated and 2 refers to the titrant.
The reduction potential of a given substance is an expression of the extent to which the
substance may take up electrons. A high positive value for the reduction potential indicates:
.
The substance is easily reduced.
The substance is a powerful oxidizing agent.
. The substance easily removes electrons from other substances with lower reduction
potential.
.
Table 5.2 lists values for reduction potentials (E0) for typical redox pairs.
Titrimetric Methods
79
Table 5.2 Reduction potentials (E0)
Oxidized form
Reduced form
E0 (V)
Ce4þ þ e
MnO4 þ 5e þ 8Hþ
Fe3þ þ e
Br2 þ 2e
I2 þ 2e
2Hþ þ 2e
Fe2þ þ 2e
Ca2þ þ 2e
Ce3þ
Mn2þ þ 4H2O
Fe2þ
2Br
2I
H2
Fe
Ca
1.61
1.51
0.55
1.05
0.54
0.00
0.44
2.89
A substance with a higher reduction potential oxidizes a compound with a lower
reduction potential. A general note for titrations is that the equilibrium constant must
be high, meaning that the reaction is complete and forms 100% product. In redox titration
the difference (DE) between the two substances reduction potentials determines the
equilibrium constant:
DE ¼ E0 T E0 A
ð5:19Þ
where E0 T is the reduction potential of the titrant and E0 A is the reduction potential of the
analyte. DE is called the reaction potential. In practice, DE should not be less than 0.1–0.2 V.
Throughout the titration the potential changes gradually with addition of titrant, and at the
Volt
1.5
Equivalence point
1.0
0.5
ml
20
40
volume of 1 M Ce 4+
Figure 5.10 Titration of 25 ml of 1.0 M Fe2þ with 1.0 M Ce4þ
80
Introduction to Pharmaceutical Chemical Analysis
E (volt)
1.2
ΔΕ = 1.0 V
1.0
ΔΕ = 0.8 V
0.8
ΔΕ = 0.6 V
0.6
ΔΕ = 0.4 V
ΔΕ = 0.2 V
0.4
0.2
ml
Figure 5.11 Titration curves with different reaction potentials
equivalence point the potential increases or decreases very sharply as a function of added
titrant. An example of this is shown in Figure 5.10 for the titration of iron (II) (Fe2þ) with
cerium (IV) (Ce4þ), which reacts according to the following equation:
Fe2þ þ Ce4þ ! Fe3þ þ Ce3þ
ð5:20Þ
For the titration of iron (II) with cerium (IV) the cell potential changes sharply at the
equivalence point because the difference in reduction potential for the Fe2þ and Ce4þ is
large. For compounds with less difference in the reduction potential the change in potential
is less pronounced as shown in Figure 5.11.
Detection of endpoint in redox titrations can be performed either by color indicators or by
potentiometry. Color indicators are compounds that are reduced or oxidized depending on
the potential and which change color from their oxidized to reduced form. Color indicators
should be selected so that the color shift is within the area where the cell potential changes
when the titration has reached the equivalence point. As for the acid–base titration, it is
important that the indicator is added in such small quantities that it does not consume
significant amounts of electrons that are exchanged. For the use of color indicators it is
normally required that the potential changes by at least 0.1 V at the equivalence point in
order to make a clear endpoint detection. In the bottom case in Figure 5.11, this change is so
small that it would be impossible to make an endpoint detection with color indicators.
Potentiometric detection can be used, and in such cases it is common to use a platinum
electrode as indicator electrode and a silver–silver chloride or a calomel electrode as
reference electrode.
In many redox titrations it is common to use sodium thiosulfate and iodine. An excess of
KI is added to the solution and the analyte oxidizes iodide to iodine by the following
reaction:
Titrimetric Methods
Analytoxidized þ 2I ! Analytreduced þ I2
81
ð5:21Þ
Excess of iodine is then titrated by sodium thiosulfate according to the following reaction:
2S2 O3 2 þ I2 ! S4 O6 2 þ 2I
ð5:22Þ
Starch is used as an indicator in the titration of iodine with sodium thiosulfate. Iodine forms a
blue-black complex with starch, and the endpoint is thus observed where the solution
changes color from blue to colorless. Sodium thiosulfate is standardized with KIO3 as a
primary standard. An exact amount of KIO3 is dissolved in an acidic solution and an excess
of KI is added. Iodine is generated by the following reaction:
IO3 þ 5I þ 6Hþ ! 3I2 þ 3H2 O
ð5:23Þ
The amount of iodine formed by this reaction is titrated with sodium thiosulfate, and the
concentration of this solution can then be determined.
Commonly used oxidizing agents are potassium permanganate, potassium dichromate
and cerium sulfate or potassium bromate. The half reactions of these are:
MnO4 þ 8Hþ þ 5e ! Mn2þ þ 4H2 O
ð5:24Þ
Cr2 O7 2 þ 14Hþ þ 6e ! 2Cr3þ þ 5H2 O
ð5:25Þ
Ce4þ þ e ! Ce3þ
ð5:26Þ
BrO3 þ 6Hþ þ 5e ! 1 =2 Br2 þ 3H2 O
ð5:27Þ
Potassium bromate is a primary standard. Potassium dichromate and cerium sulfate can act
as primary standards, but in Ph.Eur. it is customary to standardize these with sodium
thiosulfate. This also applies to potassium permanganate, which is not a primary standard.
Potassium permanganate is standardized according to the following reactions:
2MnO4 þ 10I þ 16Hþ ! 2Mn2þ þ 5I2 þ 8H2 O
ð5:28Þ
2S2 O3 2 þ I2 ! S4 O6 2 þ 2I
ð5:29Þ
A form of redox titration that is widely used in pharmaceutical analysis is the Karl Fischer
titration for precise determination of the water. That is an issue we return to in Chapter 21.
5.5 Other Principles of Titration
In addition to acid–base and redox titrations, complex reactions are also used in titration.
These are called complexometric titrations or complex formation titrations. Often it is the
concentration or amount of metal ions that is determined by complexometric titrations, and
ethylenediaminetetraacetic acid (EDTA) is often used as titrant. EDTA forms complexes
with metal ions in the ratio 1 : 1, and the endpoint of titration is usually detected by using a
color indicator.
82
Introduction to Pharmaceutical Chemical Analysis
Another titration method occasionally encountered in pharmaceutical analysis is the
Kjeldahl analysis for the determination of organically bound nitrogen. This method is based
on acid–base titration and is used extensively in connection with the total determination of
proteins. The principle of Kjeldahl analysis is that the sample analyzed is added to
concentrated sulfuric acid and a little catalyst (a Cu and Se salt). The mixture is boiled
at 340 C (boiling point of H2SO4) so that the sample is combusted. Carbon in the sample is
converted to CO2, while organic bound nitrogen is transferred to ammonium (NH4þ ). Then
NaOH is added to convert ammonium into ammonia (NH3). Ammonia is distilled into a
solution where a surplus of HCl has been added.The HCl reacts with ammonia, and the
surplus of HCl is titrated with NaOH. Based on this titration the amount of ammonia can be
determined, and from this figure the amount of organically bound nitrogen is calculated.
5.6 Summary
Titration is a key method in drug analysis and is used mainly for the quantitative determination
of active pharmaceutical ingredients (APIs) and excipients. In titration the analyte reacts in a
well defined manner with a titrant, and based on the consumption of titrant the amount of
analyte can be determined. Titrations can be performed either by acid–base reactions, by
redox reactions or by complexation reactions. The endpoint for the titration is determined
using either color indicators or potentiometry. Weak acids and bases cannot be titrated in
aqueous medium because water itself can act as an acid or base, and in this case, the titration is
performed in a non-aqueous environment.
6
Introduction to Spectroscopic Methods
This chapter presents a brief review of electromagnetic radiation and discusses how
molecules and elements absorb and emit electromagnetic radiation. Absorption and
emission of electromagnetic radiation are the basis for identification and quantitative
determinations in spectroscopic methods such as UV spectrophotometry, IR spectrophotometry, NIR spectrophotometry, atomic absorption spectrometry and atomic emission
spectrometry. These methods are presented in subsequent chapters.
6.1 Electromagnetic Radiation
Electromagnetic radiation can be described as both waves and particles called photons and
consist of regular electric and magnetic fluctuations. For many purposes it can be pictured as
an electric field that undergoes sinusoidal oscillations as it moves through space.
Figure 6.1 is a representation of a beam of monochromatic (i.e., single wavelength)
radiation. The wavelength of the sinusoidal wave can be measured as the distance between
any two points with the same phase, such as between crests, or troughs, or corresponding
zero crossings as shown in Figure 6.1.
The wavelength is commonly represented by l and is often measured in meters. In
addition to being characterized by its wavelength, electromagnetic radiation is also
characterized by its frequency. Frequency is the property of a wave that describes how
many wave patterns or cycles pass by in a period of time. Frequency is commonly
represented by n. The frequency n is measured in Hertz (Hz), where a wave with a
frequency of 1 Hz passes by at 1 cycle/s. The product of wavelength and frequency is the
speed of light (c), and it corresponds to 2.998 108 m/s in a vacuum:
ln ¼c
ð6:1Þ
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
84
Introduction to Pharmaceutical Chemical Analysis
Wavelength
Wavelength
Wavelength
Distance or time
Figure 6.1 Representation of a beam of monochromatic radiation
The energy (E) of the radiation is proportional to the frequency according the following
equation:
c
ð6:2Þ
E ¼hn ¼h
l
where h is Planck’s constant (6.626 1034 Js). This means that the energy of electromagnetic radiation increases with increasing frequency. Since frequency and wavelength
are inversely proportional, the energy of electromagnetic radiation decreases with increasing wavelength.
Various types of electromagnetic radiation are shown in Figure 6.2.
At very low wavelengths are the X-rays, the radiation of very high energy. At the opposite
end of the electromagnetic spectrum there are radio waves, which have high wavelength and
low energy. In the context of pharmaceutical analysis, only three small areas in the
electromagnetic spectrum are discussed in this book:
.
Ultraviolet radiation (UV);
Visible light (Vis);
. Infrared radiation (IR) and near-infrared radiation (NIR).
.
UV radiation is electromagnetic radiation in the wavelength region 200–400 nm
(nm ¼ nanometer ¼ 1012 m). This spectral range is used for UV spectrophotometry,
which is discussed in more detail in Chapter 7. Visible light (Vis) is electromagnetic
radiation that can be perceived by the human eye, which extends from about 400 to 800 nm.
Cosmic
rays
γ − rays
10 −12 m
X-rays
10 −11 m
UV and
visible
10 −8 m
Infrared
10 −6 m
Microwave
10 −3 m
Figure 6.2 The electromagnetic spectrum
Radio
waves
10 −1 m
Introduction to Spectroscopic Methods
Acetylsalicylic acid
A
b
s
o
r
b
a
n
c
e
85
Paracetamol
A
b
s
o
r
b
a
n
c
e
225
275
250
300
225
(nm)
250
275
300
(nm)
Figure 6.3 UV spectra of acetylsalicylic acid and paracetamol
The color orders of the visible spectrum are red, orange, yellow, green, blue, indigo and violet.
This spectral range is briefly discussed in Chapter 7 with UV spectroscopy. The area from 800
to 15 000 nm is called the infrared range, and is used in IR spectroscopy (Chapter 8).
6.2 Molecules and Electromagnetic Radiation
When electromagnetic radiation passes through molecules dissolved in a solvent such as
water the intensity of radiation is weakened. The weakening of the electromagnetic
radiation is highly dependent on wavelength. The reduction will be strong at some
wavelengths, while radiation at other wavelengths will pass largely unhindered through
the solution. When radiation weakens we say that the molecules (matter) absorb electromagnetic radiation. At some wavelengths we will thus have a strong absorption, while
absorption will be small or very small at other wavelengths. The absorption of electromagnetic radiation can be plotted as a function of the wavelength of radiation, and we get an
absorption spectrum for the particular compound. Absorption spectra in the UV range (UV
spectra) are shown for acetylsalicylic acid and paracetamol in Figure 6.3.
The wavelength where the compounds have their maximum absorbance is called the
absorption maximum. Absorption spectra are different for different chemical compounds
and can therefore be used for identification. Acetylsalicylic acid has an adsorption
maximum at 275 nm and paracetamol has an absorption maximum at 245 nm, as shown
in Figure 6.3. The amount of radiation absorbed is related to the concentration of compound
in a test solution. The absorption maximum is unaffected by the concentration of a
compound. In quantitative determinations the amount of absorbed radiation is measured
and related to concentration. The more concentrated the more radiation is absorbed.
86
Introduction to Pharmaceutical Chemical Analysis
Absorption of electromagnetic radiation of molecules forms the basis of UV, Vis, IR and
NIR spectrophotometry, to be presented in more detail in the next chapters.
Electrons in molecules are normally present in the lowest energy state, which is called the
ground state. When a photon passes near an electron, absorption becomes possible if the
energy of the photon matches exactly one of the higher energy states of the electron. The
energy of the photon is then transferred to the electron (present in an atom, ion or molecule)
converting it to a higher energy state called the excited state. Excitation of species M to its
excited state M can be described by the following equation:
M þ hn ! M*
ð6:3Þ
Excited molecules or atoms are very unstable and relax to their ground state after a very short
time (106 to 109 s). This process may also release heat to the surroundings:
M* ! M þ heat
ð6:4Þ
Relaxation may also occur by reemission of radiation, which we shall come back to later, or
by photodecomposition to form new products.
The energy absorbed can affect molecules in different ways:
.
Transfer an electron to an orbital with higher energy (electronic transition);
Increase the vibrations of the molecule (vibrational transition);
. Increase the bonding rotations of the molecule (rotational transition).
.
A particular molecule will have a large number of possible combinations of these, and each
combination represents a very specific amount of energy. This means that each possible
combination will be equal to the absorption of radiation at one specific wavelength. The
spectra of molecules consist therefore of large numbers of different absorptions. All these
absorptions, which are very close to each other, will in practice overlap and you get a
continuous spectrum as shown for acetylsalicylic acid and paracetamol in Figure 6.3.
Continuous spectra are characteristic of molecular compounds.
In some cases, parts of the energy from the excited molecules are emitted as electromagnetic radiation. This means that the sample itself emits radiation. This is called
fluorescence, and fluorescence from molecules is also a continuous radiation. Figure 6.4
shows an absorption spectrum and a fluorescence spectrum of the same compound.
Fluorescence is observed at slightly higher wavelengths than the radiation used to excite
the molecule. This is because some of the energy used for excitation will be emitted as heat.
Fluorescence will thus have a lower energy and thus a higher wavelength. Fluorescence can
be used for quantitative determinations because the fluorescence intensity is dependent on
the concentration of the fluorescent molecule. Fluorescence can also be used as a detection
method in liquid chromatography (HPLC) and will be briefly discussed in Chapter 13.
6.3 Atoms and Electromagnetic Radiation
Atomic absorption spectrometry (AAS) and atomic emission spectrometry (AES) are
presented in Chapter 9. Both AAS and AES are primarily used for quantitative determinations. Both of these techniques are based on samples being decomposed into free atoms at
Introduction to Spectroscopic Methods
Figure 6.4 Absorption and fluorescence spectra of a molecular compound
Absorbance
595
330
285
200
300
400
500
600
700
Wavelength (nm)
Figure 6.5 Absorption spectrum for sodium
87
88
Introduction to Pharmaceutical Chemical Analysis
high temperature. In AAS free atoms absorb electromagnetic radiation, and in AES free
atoms emit electromagnetic radiation when they are exposed to high temperatures. This
process is called emission.
In Section 6.2 on molecules and electromagnetic radiation it was discussed that
molecules have a very large number of ways to absorb electromagnetic radiation, and
that this results in continuous spectra. This is not the case for free atoms. Free atoms can only
absorb electromagnetic radiation by electrons being transferred to orbitals with higher
energy. Consequently free atoms have only a small number of ways to adsorb electromagnetic radiation. Each of these requires a specific energy, which results in absorption at a few
very specific wavelengths of the spectrum. This is shown in Figure 6.5 for sodium. In this
spectrum there are three major absorption lines at wavelengths 285, 330 and 590 nm. These
absorption lines are due to an electron in the 3s orbital being excited to respectively the 5p,
4p and 3p orbitals. Atomic spectra consisting of narrow absorption lines are called line
spectra. Emission of radiation from free atoms has emission lines at the same wavelengths.
6.4 Summary
Electromagnetic radiation used in pharmaceutical analysis is divided into the UV region
(200–400 nm), visible range (400–800 nm) and IR range (800–15 000 nm). Molecules
absorb electromagnetic radiation by transferring electrons to orbitals of higher energy
(electronic transitions), by vibrational transitions and by rotational transitions. Absorption
can be used both for identification and quantitative determinations. Some substances release
energy by emitting electromagnetic radiation, and this is called fluorescence. Free atoms
absorb electromagnetic radiation by transferring electrons to orbitals with higher energy,
and the excited atoms will emit radiation. Both the absorption and emission of radiation
from free atoms are used for quantitative determinations.
7
UV Spectrophotometry
This chapter reviews the principles of quantitative determination and identification based
on UV spectrophotometry. UV spectrophotometry is an official method in Ph.Eur. It is used
for quality control of active ingredients, excipients and pharmaceutical products. Instrumentation is discussed briefly and some practical tips in conjunction with UV spectrophotometry are also discussed in this chapter. Examples of how the technique is used for
identification and quantitative determinations with regard to ingredients and pharmaceutical products are discussed in Chapters 21 and 22, respectively.
7.1 Principle of Quantitative Determination
When monochromatic electromagnetic radiation (radiation with only one wavelength) with
the intensity of I0 passes through a solution of an analyte, some of the radiation is absorbed
by the analyte while the rest passes right through. When the intensity of the transmitted
monochromatic radiation, which is measured at the backside of the solution, is I (Figure 7.1),
the absorbance of radiation (A) is defined as:
1
I0
A ¼ log10
¼ log10
ð7:1Þ
T
I
where T is the transmittance (I/I0).
Beer’s law governs the measurement of absorption of radiation by a solution of
molecules. According to Beer’s law, the absorbance (A) is proportional to the path length
through which the radiation passes (b) and the concentration of the substance in the
solution (c):
A ¼ ebc
ð7:2Þ
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
90
Introduction to Pharmaceutical Chemical Analysis
Io
I
Sample solution
b
Figure 7.1 Beer’s law – absorption of radiation
When b is expressed in centimeters and c in moles per liter (M), e is called the molar
absorptivity, which is constant for that particular substance. Beer’s law states that the
absorbance is proportional to the molar absorptivity, the path length and the analyte
concentration in the solution. When e and b are known from before, the unknown molar
concentration of a drug can be determined by reading the absorbance.
The specific absorbance can be used to calculate concentrations as % (g/100 ml). The
specific absorbance is defined as:
A11 percent
¼
cm
10e
Mr
ð7:3Þ
and this refers to the absorbance of a solution containing 1 g of the substance dissolved
in 100 ml of solvent measured at a defined wavelength at a path length of 1 cm. Mr is
the relative molecular mass (molar mass). The specific absorbance is also written as A
(1%, 1 cm).
The path length is specified for the equipment used and measurements are normally done
at a path length of 1 cm. The absorbance of an analyte must be determined after calibration.
In calibration the absorbance of standard solutions of known concentrations is measured. An
example of this is shown in Box 7.1.
The absorbance is dependent on:
.
Wavelength;
Solvent (Figure 7.2);
. pH when the substance is an acid or base (Figure 7.3);
. Temperature (in relatively small extent).
.
It is important that these parameters are the same for standards and sample solution. If this is
not the case, significant errors in the quantitative determinations may occur.
When it is impossible to produce standard solutions within the same solvent and pH as in
the sample solution, calibration is slightly more complicated. This is because the absorptivity of the substance in the standard solution may be different from that in the sample
UV Spectrophotometry
91
Box 7.1 Determination of caffeine
The concentration of caffeine in an aqueous solution (sample solution) should be
determined. First a standard solution of caffeine in water with a known concentration
of 4.5 105 M is prepared. The absorbance measured at 273 nm was 0.454 with a path
length of 1 cm. From this the molar absorptivity of caffeine at the given conditions can be
calculated:
A¼ebc
or:
e ¼ A=ðb cÞ ¼ 0:454=ð1 4:5 105 Þ ¼ 10 089
The absorbance reading of the sample solution was 0.367 at the same wavelength
(273 nm), with the same path length (1 cm) and in the same solvent (water) as in the
standard solution. From this the concentration of caffeine in the sample solution can be
calculated:
A ¼ ebc
or:
c ¼ A=e b ¼ 0:367=ð10 089 1ÞM ¼ 3:64 105 M
Absorbance
Spectrum
in hexane
Spectrum
in water
450
550
650
Wavelength (nm)
Figure 7.2 Absorption of radiation is dependant on solvent. Ultraviolet and visible spectra of
tetrazole in n-hexane and in water
92
Introduction to Pharmaceutical Chemical Analysis
Absorbance
2.0
1.5
0.1 M NaOH
1.0
0.1 M HCl
0.5
220
260
300
340
Wavelength (nm)
Figure 7.3 Absorption of radiation is dependant on pH. UV spectra of phenylephrine in 0.1 M
NaOH and in 0.1 M HCl
solution. In such situations a standard addition has to be used. In a standard addition a
known amount of a chemical reference substance of the analyte is added to the sample
solution and the absorbance is measured both before and after addition of the reference
substance. Then, the concentration of analyte is calculated according to Beer’s law
[Equation (7.2)]. With a standard addition the calibration is done directly in the sample
solution to avoid any experimental variations between samples and standards of solvent, pH
and the possible presence of other substances. An example of standard addition is shown
in Box 7.2.
Beer’s law applies to monochromatic radiation, that is, radiation of one wavelength. This
is an ideal situation, which in practice cannot be achieved with the instruments used. In
practice, the radiation transmitted to the specimen form a narrow wavelength range (the
smaller the better), and consequently the sample is irradiated with polychromatic radiation.
Polychromatic radiation leads to deviation from Beer’s law. It has been shown that this
discrepancy is greatest when the value of e (absorptivity) varies greatly within the
wavelength range that hits the sample. Therefore, for quantitative measurements one
should choose a narrow wavelength range with small variations in absorptivity. This is
usually the case at the absorption maximum, which is the wavelength of maximum
absorption, as shown for band A in Figure 7.4. In addition, the best sensitivity for the
method (slope of the calibration curve) is obtained at the absorbance maxima. As shown in
Figure 7.4 the absorptivity varies greatly for band B and the sensitivity is much lower than
for band A.
In addition to exposing the sample to radiation with a narrow wavelength range, it is
important that the concentration of the analyte is not too high. In concentrated solutions the
absorbing molecules come so close together that they affect the charge distribution of each
UV Spectrophotometry
93
Box 7.2 Determination of caffeine by standard addition
The concentration (c) of caffeine in a sample solution should be determined. Because
the composition of the sample solution is unknown it is difficult to make a standard
solution of caffeine to determine the molar absorptivity. Therefore, the method of
standard addition was chosen. Two 10.00 ml portions of the sample solution is
collected with a pipette and filled into two 20.00 ml volumetric flasks. Flask A is then
filled up to the mark with water. To flask B is added 2.00 ml of a 4.5 104 M
standard solution of caffeine in water and then water is added to the mark. The
absorbance of solution A was 0.356 at 273 nm (1 cm radiation path) while the
corresponding absorbance for solution B was 0.810. For solution A, we can set
up the following formula, based on Beer’s law:
0:356 ¼ e 1 c
ð1Þ
where e is the molar absorptivity (unknown) and c is the concentration after dilution
(from 10.00 ml to 20.00 ml). For solution B, we accordingly set up:
0:810 ¼ e 1 ðc þ 4:5 105 MÞ
ð2Þ
where c is the concentration of caffeine in the sample (after dilution) while
4.5 105 M is the concentration of caffeine added as standard after dilution (from
2.00 ml to 20 ml). The two Equations (1) and (2) can be combined in the following way:
0:356=0:810 ¼ ðe 1 cÞ=½e 1 ðc þ 4:5 105 MÞ
or:
0:440 ¼ c=ðc þ 4:5 105 MÞ
or:
c ¼ 3:54 105 M
The concentration in the original sample was 2 3.54 105 M ¼ 7.08 105 M
since the sample solution was diluted by a factor of two in both flasks.
other. This will affect their ability to absorb radiation and could result in deviation from
Beer’s law. Typically, the deviation from Beer’s law is small when analyte concentrations
are below 0.01 M. For practical work, this means that relatively concentrated solutions
should be diluted before measurements are made.
A third important practical requirement for Beers Law is that the analyte should not
participate in chemical reactions in the solution. Therefore, the analyte should be stable
in the selected solvent, and the solution cannot contain substances that react with the
analyte.
94
Introduction to Pharmaceutical Chemical Analysis
Absorbance
Absorbance
A
A
B
B
Wavelength
Concentration
Figure 7.4 Calibration curves at two wavelength bands, A and B
7.2 Principle of Identification
In addition to quantitative analysis UV spectrophotometry is also used for identification of
substances. The identification is based on UV spectra, where the absorbance of the
substance is recorded as a function of wavelength. Different substances may have different
UV spectra (Figure 7.5). An unknown substance may be identified by UV spectrophotometry if the spectrum of the unknown substance is identical to a reference spectrum of a known
substance. Despite this, one should, be careful about basing identification solely on UV
spectrophotometry because different substances can have similar UV spectra. In such cases
Codeine
225
250
Ephedrine
275
nm
300
225
250
Levodopa
275
nm
300
225
250
275
nm
Figure 7.5 UV spectra of codeine, ephedrine, and levodopa
300
UV Spectrophotometry
95
it is necessary to complement the identification with other techniques such as IR
spectrophotometry.
7.3 Which Substances Have Strong UV Absorbance?
The absorption of UV radiation involves, as mentioned in Chapter 6, the transfer of electrons
to orbitals of higher energy. Two types of electrons can participate in such excitation:
.
.
Electrons shared by several atoms and which participate directly in the chemical bonding;
Unshared electrons that are localized to one atom such as O, N, S and halogens.
Electrons involved in single bonds between C and H and between C and C are held largely in
place. These electrons require a lot of energy to be excited. Absorption from these electrons
is found only at very low wavelengths in the UV range (below 200 nm), which contains a lot
of energy. For practical reasons it is not customary to measure below 200 nm, and normally
electrons in single bonds do not contribute to the absorbance measured in UV
spectrophotometry.
The absorbing groups of a molecule are called chromophores. Double or triple bonds of
organic compounds are chromophores and will contribute to the absorbance of UV
radiation. Such electrons are much “looser,” and they require less energy to be excited.
UV light with wavelengths of 200 nm and above contains enough energy to excite this type
of electrons. Typical bonds within this group include C double-bound to C, N double-bound
to N, C double-bound to O, C triple-bound to C and C triple-bound to N.
In addition to double and triple bonds, atoms with unshared electrons give important
contributions to UV-absorbing properties. This is because these electrons are “relatively
loose” and require little energy for excitation. These electrons are found for O and N and
halogens. A UV-absorbing property of a substance and the UV spectrum is therefore a result
of the sum of the various electrons that is easily excited by the UV light. When a substance
contains many double or triple bonds, conjugated double bonds, aromatic rings or contains a
number of O, N or halogen atoms, it gives a strong absorption in significant parts of the UV
range.
7.4 Instrumentation
The instruments suitable for measuring in the ultraviolet and visible range of the spectrum
(spectrophotometers) consists of an optical system capable of producing monochromatic
radiation in the range 200–800 nm and a device suitable for measuring the absorbance.
A spectrophotometer consists of a radiation source, a wavelength selector, a sample
container (called a cell or cuvette) and a detector. A schematic view of a UV spectrophotometer is shown in Figure 7.6. The radiation source emits UV radiation, the wavelength
selector selects the wavelength that irradiate sample, and the detector measures the intensity
of radiation passing through the sample solution (I). The intensity of this radiation is
compared with the intensity from the radiation source (I0), and converted by an electronic
system to absorbance units [log (I0/I)] displayed on the instrument.
96
Introduction to Pharmaceutical Chemical Analysis
Sample solution
Wavelength
selector
Detector
Lamp
Figure 7.6 Schematic view of a UV spectrophotometer
Radiation sources used in modern spectrophotometers are continuous sources that emit
radiation of all wavelengths within a given area. Because Beer’s law applies to monochromatic radiation, it is important to have a wavelength selector before the sample container.
A small wavelength range will then be directed from the radiation source and on to the
sample cell. Deuterium lamps are normally used as radiation source in the UV range. These
lamps emit continuous radiation in the range 160–375 nm. Many spectrophotometers also
cover the area of visible radiation. In the visible range tungsten lamps that emit continuous
radiation in the range 350–2500 nm are used. To achieve stable conditions during absorbance measurements it is important that the radiation does not fluctuate, that is, the intensity
of radiation should remain constant over time. This makes great demands on both the lamps
and the corresponding voltage sources, and the lamps are relatively costly.
The wavelength selector is normally placed between the radiation source and the sample
container and allows only radiation with a narrow wavelength range to pass from the lamp to
the sample. Normally a monochromator is used as wavelength selector. Two common
monochromators used in UV/Vis spectrophotometers are prisms and reflection gratings.
The outline of a monochromator based on a reflection grating is shown in Figure 7.7.
The monochromator is used to disperse the light into its constituent wavelengths. Continuous radiation is transmitted through a small slit (entrance slit) and towards a concave mirror.
The radiation reflected by the mirror is sent on to a small plate called a reflection grating.
This is a hard and flat thin plate with a large number of small parallel grooves in a reflective
coating. For the UVand the visible region it is typically between 300 and 2000 grooves/mm,
with 1200–1400 being most common. Continuous radiation that hits the grating will be
reflected and transmitted at various angles depending on the wavelength. The radiation from
the grating hits a new concave mirror and is passed on to a new slit (output slit). A given
small wavelength range will pass through the slit and on to the sample, while all other
wavelengths (one such visualized with dotted line) are reflected in different angles from the
grating and the mirror and cannot pass through the output slit. The wavelength of the beam
that irradiates the sample can be varied continuously by varying the position (angle) of the
reflection grating. When the grating is turned slightly, another small wavelength range will
UV Spectrophotometry
97
Concave mirror
Concave mirror
Reflection grating
Output
slit
Input
slit
Polychromatic
radiation
Monochromatic
radiation
Figure 7.7 Schematic view of a monochromator based on a reflection grating
pass through the output slit and in this way another wavelength can be selected. The
monochromator can be used both in the UV and the visible region.
Some newer instruments use a photo diode array (Figure 7.8). Continuous radiation is
transmitted through the sample and hit the reflection grating. This reflects radiation in
different angles according to wavelength as previously discussed. The reflected radiation
strikes the photo diode array.
A photo diode array is an array of a large number of small photo diodes that detect the
intensity of the radiation that hits them. Each diode measures the radiation in a given
wavelength interval. This means that the absorption of the light can be detected and amplified
to give an absorption spectrum or a response on one wavelength or different wavelengths.
The samples to be analyzed by conventional UV spectrophotometry are always dissolved
in a suitable solvent prior to measurement. The sample solution is placed in a sample
container (cell or cuvette) that is irradiated in the spectrophotometer. The sample cells must
have windows fabricated from a material that is transparent to the radiation used. They are
typically small containers of quartz, glass or plastic, often made with a square cross-section
with an inner width of exactly 1.0 cm. They can be supplied with a cap for analysis of
samples made in volatile solvents. The cap prevents evaporation during measurement, and
hence eliminates fluctuations in the concentration that may cause significant errors in
quantitative determinations. Sample containers of glass or plastic can be used in the visible
range and are usually relatively cheap. For measurements in the UV region neither glass nor
98
Introduction to Pharmaceutical Chemical Analysis
Figure 7.8 A simplified diode array configuration
plastic can be used because they absorb electromagnetic radiation in the UV range. In such
cases, one should use sample containers made of quartz. In the spectrophotometer the
intensity of radiation that irradiates the sample cell (I0) and the radiation transmitted through
the cell (I) are measured by a detector. The electronic system in the instruments calculates
the absorbance [log (I0/I)], which is shown on the instrumental display.
Interaction between the radiation and the walls of sample container is inevitable; with
losses in power occurring at each interface as the result of reflection and possibly absorption.
In order to compensate for these effects, the intensity of the radiation transmitted through the
sample cell containing an absorbing analyte is generally compared with the intensity of
radiation that passes through an identical cell containing only solvent. UV spectrophotometers can be built up either by the single-beam principle, or by the double-beam principle
(Figure 7.9). Single-beam instruments have only one sample cell. This means that the
absorbance of a solution that contains only solvent and the sample solution cannot
be measured simultaneously. First, the solvent most be placed into the sample cell and
Single-beam instrument
Monochromator
UV lamp
Detector
Reference solution/sample solution
Double-beam instrument
Sample
solution
Monochromator
UV lamp
Detector
Reference
solution
Figure 7.9 Single-beam and double-beam spectrophotometers
UV Spectrophotometry
99
the absorbance of the solvent must be adjusted to zero. Then that cell is removed and a cell
with the sample solution is inserted and the absorbance is measured. As the absorbance of
the solvent and the sample solution are not measured simultaneously inaccuracies may
occur since the characteristics of the radiation source, the detector and the electronics may
change over time. This has been taken into account in the double-beam instruments, where
the electromagnetic radiation from the radiation source is split into two parallel beams. One
beam passes through a cell for the solvent (reference cell), while the other beam passes
through another cell with the sample solution (sample cell). With this configuration, the
absorbances of the reference cell and the sample cell are measured simultaneously. Doublebeam instruments thus compensate for variations in the characteristics of the radiation
source, detector and the electronic system. This means that measurements can be done faster
and with greater accuracy.
7.5 Practical Work and Method Development
In the development and use of methods based on UV spectrophotometry, the following
points are essential:
1. Selection of solvent used to dissolve the sample.
2. Selection of wavelength used to read the absorbance.
3. Calibration and preparation of standards.
4. Handling of samples in the UV instrument.
When analysis is performed according to monographs in Ph.Eur. points 1–3 are given in the
monographs, and only point 4 requires special attention by the operator. However, when
developing a method based on UV spectrophotometry all points must be considered.
When UV spectrophotometry is used, the first step in method development is to find a
suitable solvent. It is essential that the solvent dissolves the analyte and that particles from
other undissolved substances in the sample is removed prior to absorbance measurement.
Particles in the solution may case light scattering and reflection of the radiation and
inaccuracies in the absorbance reading. The main bulk of active pharmaceutical ingredients
are organic compounds with basic functional groups and these will usually be well soluble in
either acidic aqueous solutions (e.g., hydrochloric acid) or in organic solvents such as
ethanol or methanol. The solvent used should be transparent to the radiation at the
wavelength range used (Table 7.1).
Table 7.1 Solvents for UV spectrophotometry
Solvent
Water
Acetonitrile
n-Hexane
Methanol
Ethanol
Cyclohexane
Cut-off (nm)
Solvent
Cut-off (nm)
190
190
195
205
210
210
Diethyl ether
1,4-Dioxane
Dichloromethane
Chloroform
benzene
Acetone
210
220
220
240
280
330
100
Introduction to Pharmaceutical Chemical Analysis
Absorbance
Absorbance
290 nm
290 nm
410 nm
410 nm
Wavelength
Concentration
Figure 7.10 Calibration curves at different wavelengths
All solvents absorb radiation at lower wavelengths. At higher wavelengths, most solvents
are transparent (do not absorb radiation). Above this wavelength, which is called the cut-off,
the solvents can be used for UV spectrophotometry. Water and acetonitrile have a low UV
cut-off and are used as solvents above 190 nm, while acetone has a high UV cut-off and can
only be used above 330 nm. The final decision should also include a verification of the
stability of the analyte in the solvent. The solvent should not react with the analyte or
catalyze decomposition reactions. Usually this is not a problem, but analyte stability should
be verified experimentally prior to the final decision.
The absorbance readings should preferably be at the wavelength where the analytes
have their absorption maxima. Quantitative measurements should be made at this
wavelength for maximum sensitivity (Figure 7.10) and to achieve the best possible linear
calibration curves (Figure 7.4).
At wavelengths where the analyte has a strong absorbance the calibration curve is steep,
and two samples with little difference in concentration result in significantly different
absorbance readings. In this case, the sensitivity, which is defined as the slope of the
calibration curve, is high. At wavelengths where the analyte has a low absorbance, two
samples with significantly different concentrations may give similar absorbance readings.
In the latter case, the slope of the calibration curve is small and the sensitivity is low.
Published values for absorptivity should be used with great caution because these depend
on the experimental conditions. When developing new UV methods, it is normally advised
to set up a calibration curve (absorbance as a function of concentration) based on chemical
reference standards. Standard solutions must be prepared in the same way as the sample, so
that the solvent, pH and possible presence of other substances are the same for sample and
standards. The calibration curve should be set up in the absorbance range 0.2–0.8 because
this range offers the greatest precision. The samples must be diluted accordingly if the
absorbance readings are above 0.8 absorbance units.
UV Spectrophotometry
101
When samples and standards are dissolved in organic solvents, it is important that all
sample containers are closed with a cap to prevent evaporation. Especially from the sample
cells evaporation can cause significant changes of concentrations because the volume of
these is usually relatively small (4–5 ml). Even with solvents like ethanol evaporation is
significant and can cause problems in quantitative determinations.
It is of particular importance that sample cells made of quartz are clean to achieve the
best possible optical quality. The walls should not be touched because impurities on the
walls such as the grease stains, etc., may yield significant errors. Before absorbance
readings the cells should be wiped with lens paper soaked in methanol or ethanol. Drying
of the cells in flames or ovens cannot be recommended because this can change their
optical properties.
7.6 Areas of Usage and Performance
UV spectrophotometry is particularly suitable for quality control of active ingredients
and excipients (see Chapter 21) and for quality control of simple pharmaceutical
products (see Chapter 22). For more complex samples, the method is not used directly,
but UV spectrophotometry combined with liquid chromatography (HPLC) is a standard
method for the analysis of pharmaceutical products (Chapter 13). In this case the HPLC
system comprises a UV detector (HPLC-UV). UV detection is also used in capillary
electrophoresis and in dissolution testing. In addition, HPLC-UV is used for the
determination of drugs in blood and urine (Chapter 23). For UV spectrophotometry
to be applicable the analyte must have UV absorbance or it must be converted to a
substance that absorbs UV light.
Compared with many other analytical methods UV spectrophotometry is rapid and easy
to perform. Although the method requires a UV spectrophotometer, the instrument is
relatively cheap in comparison with other instrumentation used in pharmaceutical analysis.
The cost of using UV spectrophotometry is therefore relatively low.
In addition, the accuracy and the precision are important parameters in evaluation of a
method. Accuracy indicates the difference between the experimentally determined values
and the true value. The official UV methods for purity control of raw materials in Ph.Eur. are
tested in a number of laboratories, and the accuracy for these are usually better than 1–3%.
Similar results are also found for UV methods used for pharmaceutical products.
The precision of an analysis indicates the difference between multiple measurements
(parallels) of the same sample. UV spectrophotometry provides a high degree of precision,
but the precision is normally not as high as that obtained with titration.
7.7 System Testing
When using spectrophotometers, it is essential that the equipment is calibrated to ensure
correct wavelength and absorbance readings. According to Ph.Eur., the wavelength scale is
verified using the absorption maxima of holmium perchlorate solution R. The absorption
maxima read on the instrument should be as specified in Table 7.2. The permitted tolerance
is 1 nm for the ultraviolet range and 3 nm for the visible range.
102
Introduction to Pharmaceutical Chemical Analysis
Table 7.2 Absorption maxima for control of wavelength scale
Maxima (nm)
241.15
287.15
361.5
536.3
Table 7.3 Specific absorbance of potassium dichromate for control of absorbance
Wavelength (nm)
235
257
313
350
430
Specific absorbance (A1%, 1 cm)
Maximum tolerance
124.5
144.5
48.6
107.3
15.9
122.9–126.2
142.8–146.2
47.0–50.3
105.6–109.0
15.7–16.1
In addition, the absorbance scale should be calibrated regularly. For control of absorbance
according to Ph.Eur., solutions of known concentrations of dried potassium dichromate R
are prepared. The absorbance is read at the wavelengths indicated in Table 7.3, which gives
for each wavelength the exact value and the permitted limits of the specific absorbance.
The value of the specific absorbance should be within the specified limits.
7.8 Summary
UV spectrophotometry is a key method in pharmaceutical analysis and is used both for
identification of raw materials and pharmaceutical products and for quantitative determinations. A beam of UV light is directed through the sample cell. In quantitative determinations
the absorbance is read at a given wavelength. Based on Beer’s law the absorbance is
converted to a concentration. The identification is based on absorbance spectra and
substances are identified by comparison of their absorbance spectra with reference spectra.
8
IR Spectrophotometry
This chapter reviews infrared(IR) spectrophotometry and the closely related techniquenear infrared(NIR) spectrophotometry. Both techniques are official methods in Ph.Eur.
IR spectrophotometry is primarily used for identification of pharmaceutical ingredients
and pharmaceutical products. The chapter discussed the information that can be obtained
in the different parts of the IR spectra and the instrumentation used. Examples of how the
technique is employed for identification of raw materials and products are discussed in
Chapters 21 and 22. NIR spectrophotometry is used both for quality control of pharmaceutical ingredient, for process control and for the control of finished products. The
information that can be obtained from NIR spectra and the instrumentation used are
discussed briefly.
8.1 IR Spectrophotometry
IR spectrophotometry is based on the fact that all atoms in organic molecules are in
continuous vibration and rotation with respect to each other. When the frequency of a
specific vibration is equal to the frequency of the IR radiation directed to the molecule, the
molecule absorbs the radiation and the radiation is converted into energy of molecular
vibration.
IR spectroscopy follows the same basic principles for absorption of electromagnetic
radiation as described for UV spectroscopy (Figure 8.1). Normally the wavelength region
between 2500 and 15 000 nm is of greatest practical use. This region is called the middle
infrared region and is discussed further in this chapter. In recent years, the range of
800–2500 nm has also been used in pharmaceutical analysis. This region is called the near
infrared region (NIR) and is utilized in NIR spectroscopy, which is discussed at the end of
this chapter.
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
104
Introduction to Pharmaceutical Chemical Analysis
IR source
Sample
Detector
Figure 8.1 Basic principles of IR spectroscopy
IR spectra show more detailed absorption bands than UV spectra of the same compound.
IR spectra are like fingerprints of organic molecules and provide valuable information used
to identify compounds. This is demonstrated in the UV and IR spectra of paracetamol in
Figure 8.2.
A principal difference between UV spectra and IR spectra, is that in UV spectra
absorbance is plotted as a function of wavelength and in IR spectra transmittance is plotted
as a function of wavenumber (see Figure 8.2). Transmittance is defined by:
T ¼ I=I0
ð8:1Þ
where I0 is the intensity of radiation directed to the sample and I is the intensity of radiation
transmitted through the sample. Transmittance is often expressed as a percentage, using the
following formula:
% T ¼ I=I0 100%
ð8:2Þ
When the sample does not absorb radiation, I ¼ I0 and the transmittance is 100%. If all
radiation is absorbed by the sample, the transmittance is 0%. The transmittance in IR
spectra is therefore scaled from 0 to 100%. Wavenumber is usually used instead of
wavelength on the other axis. Wavenumbers are defined by:
Wavenumber ¼ 1=wavelength ðin cmÞ
Absorbance
ð8:3Þ
Transmittance
UV
IR
225
300
Wavelength (nm)
2000
650
Wavenumber (cm -1)
Figure 8.2 UV and IR spectra of paracetamol
IR Spectrophotometry
105
High wavenumbers correspond to low wavelengths and low wavenumbers correspond to
higher wavelengths. Wavenumbers are usually expressed as cm1, and the most common
spectral range is the wavelength region between 2500 and 15 000 nm corresponding to
wavenumbers between 4000 and 670 cm1.
IR spectra can roughly be divided into two regions:
Region for reading of functional groups (1200–3600 cm1)
. Fingerprint region (600–1200 cm1)
.
The first-mentioned region is normally used to determine functional groups and bonds of a
molecule. This region is relevant when the structure of the analyte is completely unknown.
In identification work according to pharmacopeias this area has limited value, as the
fingerprint region is more useful for the identification of compounds when reference spectra
are available. IR spectra usually provide detailed information in the fingerprint region, and
the agreement between a spectrum of an unknown substance and a reference spectrum for a
known substance provides a highly secure identification (Table 8.1).
The spectral range of functional groups is typically from 1200 to 3600 cm1. As shown in
Table 8.1 different functional groups and bonds have characteristic absorption bands in the
IR spectrum, which are independent of other functional groups and bonds. Figure 8.3 shows
an IR spectrum of salicylic acid. The absorption band at 3220 cm1 is due to the phenol
group. At 3050 cm1 there is a relatively weak absorption band derived from the CH bonds.
At 1650 cm1 there is a strong absorption band from the C¼O function in the acid group
and the absorption band at 1140 cm1 is due to C¼C bonds in the aromatic ring.
Interpretation of IR spectra is not discussed further as this is mainly used in structure
elucidation of unknown compounds.
Table 8.1 Some characteristic infrared absorption peaks
Binding Functional group
CH
OH
NH
C¼C
CN
CO
C¼O
NO2
Alkanes
Wavenumber (cm1) Intensity
2850–2970
1340–1470
Alkanes
3010–3095
675–995
Alkynes
3300
Aromatic rings
3010–3100
690–900
Alcohols, phenols
3650–3950
H-bound to alcohols, phenols
3200–3600
Carboxylic acids
3500–3650
H-bound to carboxylic acids
2500–2700
Amines, amides
3300–3500
Alkenes
1610–1680
Aromatic rings
1500–1600
Amines, amides
1180–1360
Alcohols, ethers, esters, carboxylic acids
1050–1300
Alcohols, ketones, esters, carboxylic acids 1690–1760
Nitro substituents
1500–1570
1300–1370
Strong
Strong
Medium
Strong
Strong
Medium
Strong
Variable
Variable (broad)
Medium
Broad
Medium
Variable
Variable
Strong
Strong
Strong
Strong
Strong
106
Introduction to Pharmaceutical Chemical Analysis
Transmittance (%)
80
60
40
20
0
4000
3000
2000
1500
1000
600
Wavelnumber (cm -1)
Figure 8.3 IR spectrum of salicylic acid
The fingerprint region normally extends between 600 and 1200 cm1. Many absorption
bands that may be difficult to correlate to specific functional groups or bonds characterize
this area. In spite of this, the fingerprint area is commonly used because spectra in this region
vary considerably, even for compounds with small differences in molecular structure. Since
the absorption bands are difficult to interpret in the fingerprint region, the major usage is for
identification when reference spectra are available or recorded for comparison.
IR spectra are normally recorded using samples in their solid form. Such spectra are
largely unaffected by the experimental conditions, in contrast to UV spectra recorded in
solution. Comparison with literature spectra is considerably easier in IR spectroscopy than
in UV spectroscopy. Advanced IR spectrophotometers are supplied with a library of IR
spectra that are stored on a computer. The IR spectra of unknown compounds can thus be
compared electronically with reference spectra. When the library spectra are consistent with
the spectra of unknown substances, these can be identified automatically. Libraries can be
provided with more than 100 000 different IR spectra, which are of great value for
identification work.
8.2 Instrumentation
A principal view of a double-beam IR spectrophotometer is shown in Figure 8.4.
The major components are a radiation source, two sample holders, a rotating mirror, a
monochromator and a radiation detector. The radiation source emits continuous IR
radiation that is split into two equivalent beams that pass through the sample holder and
the reference holder, respectively. The reference holder is usually empty and the sample is
placed in the sample holder. Using an optical chopper (such as a rotating mirror) the sample
beam and the reference beam are alternately directed through the monochromator.
The monochromator selects the wavelengths of the two beams that can be monitored in
turn by the detector across the range of the spectrum. The two signals are amplified and then
compared electronically for recording of transmittance. By continuously varying the
IR Spectrophotometry
Sample holder
107
Mirror
Mirror
Radiation
source
Mirror
Mirror
Monochromator
Detector
Rotating mirror
Reference holder
Mirror
Figure 8.4 Schematics of a double-beam IR spectrophotometer
wavelength of radiation entering the detector a printout of the IR spectrum is obtained. This
“double-beam” setup gives accurate spectra even if the intensity of the light source drifts
over time.
Compared with UV spectrometers the IR spectrometers differ on two essential points:
.
.
The monochromator in IR spectrometers is always placed behind the sample holder.
IR spectrometers are normally based on the double-beam principle.
As in UV spectrometers, the double-beam principle in IR spectrometers is used to correct for
variations in the intensity of the radiation source and to correct for operations in the detector
and the electronic systems. This is particularly important in the IR region because the
intensity of radiation sources is relatively low. In addition, the dual-beam principle is
important for IR spectroscopy because water vapor and carbon dioxide in air have a strong
absorption of infrared radiation (Figure 8.5). Without the dual beam configuration the IR
spectra would show characteristic absorption bands from both water and carbon dioxide,
particularly in the area between 1400 and 4000 cm1. If an IR spectrum for an organic
compound is recorded without any background correction, some of the absorption bands for
that substance could coincide with absorption bands for water and carbon dioxide. A dualbeam instrument with air in the reference holder effectively corrects the background signal
by subtracting the spectra of the two channels.
Figure 8.5 IR spectra of air recorded on a single-beam and a double-beam instrument
108
Introduction to Pharmaceutical Chemical Analysis
The common radiation source for the IR spectrometer is an inert solid heated electrically
to 1000–1800 C. Three popular sources are the Nertnst glower, Globar source and
Nichrome coil. In the Nernst glower source a small cylinder of zirconium oxide or yttrium
oxide is heated. The Globar source is based on silicon carbide. They all produce radiation in
the wavelength range 1100–40 000 nm but with different energy profiles.
Monochromators used in IR instruments are constructed in the same way as monochromators used in UV instruments. The major difference is the number of grooves per
millimeter used in the optical grating. The grating used in the UV has 300–2000 grooves/
mm, while the grating used in IR has ca. 100 grooves/mm due to higher wavelengths used in
IR spectroscopy.
The instrument described above is a dispersive spectrometer as the beams from the
radiation source are dispersed by the monochromator into component frequencies before the
beam falls on the detector. In Fourier transform IR (FTIR) spectrometers the monochromator is replaced by an interferometer. The interferometer produces interference signals
(interferogram) which contain infrared spectral information generated after passage
through the sample. A mathematical operation known as a Fourier transformation converts
the interferogram to the final IR spectrum. For many applications FTIR spectrophotometers
have replaced dispersive instruments due to their superior speed and sensitivity.
UV/Vis spectra are recorded on diluted solutions of the analyte. In IR spectroscopy, this
procedure is more complicated because most solvents (including water) absorb infrared
radiation. Spectra recorded in solution will therefore contain absorption bands from both the
analyte and the solvent. This complicates the interpretation of spectra. Sample processing
for IR spectroscopy is therefore different from that in UV/Vis spectroscopy.
The double-beam instrument allows correction of background reading from water and
carbon dioxide.
Most substances of pharmaceutical interest are solids and IR spectra of solids are
normally recorded on KBr discs. According to Ph.Eur., a small amount of the substance
(1–2 mg) is finely ground and mixed with approximately 300–400 mg of dry potassium
bromide (KBr) or potassium chloride (KCl) in a mortar. If the substance is a hydrochloride,
it is recommended to use potassium chloride. The mixture is then compressed into a
transparent disc in an evacuable die at sufficiently high pressure. The disc is then placed in
the sample holder placed between the radiation source and the monochromator. Potassium
bromide is transparent to IR radiation in the range above 400 cm1 and has no absorption
bands in the region traditionally used for IR spectroscopy. The absorption bands recorded on
the KBr disc thus originate from the analyte or impurities present in the KBr mixture. The
best results are obtained when the KBr mixture is compressed in vacuo to exclude air and
water. Residual water in the disc has absorption bands at 3450 and 1640 cm1. Too large
particles in the disc contribute to band distortion due to scattering of the radiation. In order to
minimize the effect of scattering, solid samples are ground to small particles of 2 mm or less
in size (the low end of the wavelength scale).
Newer IR instruments have the possibility to record spectra directly on the substance to be
investigated. The substance is placed in close contact with an internal reflection material
such as diamond, germanium zinc selenide or other suitable material having a high
refractive index and which do not absorb infrared radiation. A close contact between
the substance and the whole crystal surface of the internal reflection element is obtained by
pressure or by evaporation of a solution of the substance on the reflection element. The IR
IR Spectrophotometry
109
spectrum is now obtained by recording the reflection (single or multible) of the IR radiation
directed onto the substance.
Diffuse reflection is a widely used sampling method for NIR spectrophotometry, which is
discussed later in this chapter. Samples to be run by diffuse reflection are generally ground to
a fine powder and mixed with an IR transparent salt such as potassium bromide prior to
sampling.
8.3 Scope
The primary usage of IR spectrophotometry in pharmaceutical analysis is the identification
of active pharmaceutical ingredients and excipients according to monographs in Ph.Eur. IR
spectrophotometry can also be used to access the purity of a compound, but in pharmaceutical analysis other techniques are more useful. IR spectra are quick and easy to run. There
are few limitations when it comes to organic substances that can be identified by IR
spectroscopy since all organic compounds absorb infrared radiation. The usefulness of the
technique in identification work is largely determined by whether there are other substances
in the sample that will interfere in the analysis. In this case it may be difficult to use IR
spectroscopy for identification without a prior sample preparation. Therefore, IR spectroscopy is most commonly used in the identification of pharmaceutical raw materials (see
Chapter 21) and in the identification of active substances in the simpler pharmaceutical
preparations (see Chapter 22).
8.4 Instrument Calibration
IR spectrometers should be calibrated regularly to verify that absorption bands are recorded
at the correct wavenumber and to control the spectral resolution. According to Ph.Eur. the
wavenumber scale is verified using a polystyrene film, which has transmission minima
(absorption maxima) at the wavenumbers (in cm1) shown in Table 8.2. As shown in the
table, there are a number of absorption maxima for polystyrene that are used to control the
wavenumber scale. Satisfactory accuracy is obtained when the various readings are within
1.0 or 2.0 cm1.
Control of resolution performance is based on a recorded spectrum of a polystyrene film.
If resolution is too low, details in the spectra are lost, and consequently the value of the
spectral information is reduced. An example of a resolution performance test is shown in
Box 8.1.
Table 8.2 Transmission minima (cm1) and acceptance tolerances of a polystyrene film
Transmission minimum
(and acceptance tolerance)
3060.0 ( 1.5)
2849.5 ( 2.0)
1942.9 ( 1.5)
1601.2 ( 1.0)
Transmission minimum
(and acceptance tolerance)
1583.0 ( 1.0)
1154.5 ( 1.0)
1028.3 ( 1.0)
110
Introduction to Pharmaceutical Chemical Analysis
Box 8.1 Control of resolution performance
Transmission %
Record the spectrum of a polystyrene film approximately 35 mm in thickness. The figure
below shows the difference x between the percentage transmittances at the transmission
maximum A at 2870 cm1 and at the first transmission minimum B at 2849.5 cm1 must
be greater than 18. The difference y between the percentage transmittances at transmission maximum C at 1589 cm1 and at the transmission minimum D at 1583 cm1 must
be greater than 10.
80
80
60
60
40
40
C
y
A
D
x
20
20
B
3200 3000 2800 2600
1800
1600
1400
Wavenumber cm -1
8.5 NIR Spectrophotometry
Conventional IR spectroscopy as described in the previous section is normally carried out in
the transmission mode based on the absorption of infrared radiation transmitted through the
sample. In addition to the transmission mode, NIR spectroscopy is also carried out in the
diffuse reflection mode and the transflection mode as shown in Figure 8.6. The diffuse
reflection mode measures reflectance and the transflection mode measures a combination of
transmittance and reflectance. The transmission mode and the transflection mode are most
useful for liquid and semisolid samples and the diffuse reflection mode is most useful for
solid samples. As most samples of pharmaceutical interest are solid samples, the diffuse
reflection mode is the most useful for pharmaceutical analysis. It can be used directly
IR Spectrophotometry
111
TRANSMISSION MODE
NIR detector
Radiation source
Sample
TRANSFLECTION MODE
Sample
Radiation source
NIR detector
DIFFUSE REFLECTION MODE
NIR detector
Radiation source
Sample
Figure 8.6 Illustration of transmission mode, transflection mode and diffuse reflection mode
without any sample treatment. NIR spectra are recorded rapidly and NIR therefore has the
potential to produce great savings in time. A beam of NIR radiation is directed toward the
sample at a given angle, and the radiation reflected from the sample is recorded. At
wavelengths where the sample strongly absorbs radiation the detector can detect only a
small amount of reflectance. Similarly, at wavelengths where the sample has a weak
absorption a strong reflectance is recorded.
Reflectance (R) is measured as the ratio of the intensity of radiation reflected from the
sample (I) to that reflected from a background or a reference reflective surface (Ir):
R¼
I
Ir
ð8:4Þ
In NIR spectroscopy, it is common to convert the intensity of the reflectance, R, into
absorbance units AR:
1
Ir
¼ log10
AR ¼ log10
ð8:4Þ
R
I
In NIR spectra, absorbance units (AR) are plotted as a function of wavelength (nanometer ¼
nm) as shown in Figure 8.7. The spectra have only a few significant peaks, but they are
exceptionally information-rich due to the number of overlapping absorption bands. NIR
112
Introduction to Pharmaceutical Chemical Analysis
Absorbance (AR)
0.5
0.4
0.3
0.2
0.1
0
1100
1300
1500
1700
1900
2100
wavelength (nm)
Figure 8.7 NIR spectrum of an active ingredient
spectra are unique “reproductions” of a substance or a mixture. It is difficult to assign
specific features to specific chemical components. Direct comparison of the spectrum
obtained with a reference spectrum of a chemical reference substance, as used in infrared
absorption spectrometry, is not appropriate. Suitable validated mathematical treatment of
the data is required. This involves the careful development of a set of calibration samples and
the application of advanced software (for multivariate calibration) and computing facilities
to extract the desired information.
A simplified illustration of a NIR instrument for measuring diffuse reflection is shown in
Figure 8.8. The sample is filled into a sample vial and the vial is placed on a measurement
window. Radiation from the radiation source is dispersed by a monchromator and directed to
the sample. Radiation reflected from the sample is recorded by the detector. Halogen lamps
that emit radiation from about 800 to about 2500 nm are often used as radiation source. The
monochromator can be a prism, grating or an interferometer.
8.6 Applications
NIR spectrometry has a wide variety of applications for chemical analysis such as:
.
Identification of active substances, excipients, dosage forms and manufacturing
intermediates;
. Quantification of active substances and excipients;
. Process control.
IR Spectrophotometry
113
Sample vial
Sample
Measurement window
Mirror
Detectors
Grating
Radiation source
Figure 8.8 Illustration of a NIR spectrometer for measurements of diffuse reflection
Also physical properties of solid samples can be examined such as:
.
.
Particle size, crystalline form and polymorphism;
Hardness, dissolution behavior and disintegration pattern.
The analysis is based on the establishment of a spectral reference library. First, a suitable
number of batches of the substance, which exhibit the variation typical for that substance,
are fully tested according to established specifications. NIR spectra of these batches are
then recorded to establish the reference library, which represents the information needed
for identification and characterization of the substance. The similarity border for the
substance is defined and is the entry for that substance in the library. Comparison of the
spectrum of the substance under investigation is made with the individual or mean
reference spectra of all substances in the library on the basis of their mathematical
correlation. When the NIR spectrum of a test substance is within the similarity border for
that substance the quality of the substance should be within the established specifications
(see Figure 8.9).
As NIR spectra are recorded very rapidly and solid samples such as powders or solid
dosage forms are placed directly in vials in the instrument, NIR spectroscopy saves time and
lends itself very well to process control. The technique plays an increasing role in the
manufacturing of pharmaceuticals. Huge databases can be built to correlate process
efficiency to product quality.
114
Introduction to Pharmaceutical Chemical Analysis
Absorbance (AR)
0.5
0.4
0.3
0.2
0.1
0
1100
1300
1500
1700
1900
2100
wavelength (nm)
Figure 8.9 NIR spectra of samples with a similarity border within the established specifications
8.7 Summary
IR spectroscopy is a key method for the identification of pharmaceutical raw materials and
products. Infrared radiation is directed through a KBr tablet of the solid substance or directly
on powder, and the IR spectrum is recorded by plotting transmittance as a function of
wavenumber. IR spectra are quick and easy to run. Comparing the recorded IR spectra with
reference spectra of the same compound identifies the substances.
NIR spectroscopy is a relatively new technique in pharmaceutical analysis. Normally, no
sample preparation is required and the potential of the technique is rapid identification,
quantitative determination and characterization of active ingredients, excipients and dosage
forms. The establishment of spectral libraries is necessary for each substance being
examined and advanced software and computing facilities are necessary to extract the
desired chemical or physical information.
9
Atomic Spectrometry
This chapter provides general information and defines the procedures used in element
determinations by atomic absorption spectrometry (AAS) and atomic emission spectrometry
(AES). AAS and AES are official methods of Ph.Eur. and both are used for the quality control
of raw materials and products. The instrumentation used and some practical tips in
connection with quantification of elements by atomic spectroscopy are also briefly discussed
in this chapter.
9.1 Atomic Absorption Spectrometry
Atomic absorption spectrometry (AAS) is a technique used to determine the concentration
of a metallic element in a sample. Spectroscopic studies of atoms can only be performed in a
gaseous medium and an atomic vapor has to be generated from the sample. The absorption
of electromagnetic radiation by the atomic vapor is measured. AAS determines the total
concentration of an element regardless of how these elements are combined and it cannot
differentiate between elements bound to different molecules in a sample. In contrast,
molecular spectroscopy gives qualitative and quantitative information about molecules in
the sample.
The most important step in all atomic spectroscopic procedures is atomization, a process
in which the sample is volatilized and decomposed in such a way as to produce the atomic
vapor. Liquid samples are aspirated into the flame and because of the high temperature in the
flame the molecular bonds are broken, and the sample is broken into free atoms. Radiation
from the lamp is passed through the flame and volatilized atoms absorb radiation of a
specific wavelength with an energy corresponding to the difference between the ground
state and the excited state. The flame itself emits radiation at several different wavelengths,
and this must be removed so as not to interfere with the absorbance measurements.
The monochromator selects radiation of the specific wavelength emitted by the radiation
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
116
Introduction to Pharmaceutical Chemical Analysis
b
I0
I
Detector
Radiation source
Flame
Figure 9.1 Absorption of radiation in atomic absorption spectrometry
source after it has passed through the flame and directs the radiation to the detector as shown
in Figure 9.1. The detector measures the non-absorbed radiation.
When the intensity of radiation directed to the flame is I0 and the intensity of radiation
transmitted through the flame is I, Beer’s law applies:
A ¼ logðI0 =IÞ ¼ a b c
where a is the absorption coefficient of the element at the appropriate wavelength, b is
the path length through the flame and c is the total concentration of the element in the
test solution.
In direct calibration according to Ph.Eur., the absorbance (A) of not fewer than three
reference solutions of known concentrations is recorded. Their concentrations should span
the expected value of the test solution. For assay purposes, the optimal calibration levels are
between 0.7 and 1.3 times the expected content of the element to de determined or the limit
prescribed in the monograph. Each solution is introduced into the instrument using the same
number of replicates for each of the solutions to obtain a steady reading. A calibration curve
is prepared from the means of the readings obtained with the reference solutions by plotting
the mean absorbance as a function of element concentration (Figure 9.2). The concentration
of the element in the test solution is determined from the curve.
Absorbance
Concentration
Figure 9.2 Calibration curve for direct calibration
Atomic Spectrometry
117
Standard addition is an alternative to direct calibration. By standard addition the
calibration is done directly in the test solution to avoid any experimental variations between
samples and standards. By atomization in flames, processes can to some extent be influenced
by the viscosity and the content of other components in the test solution and in these cases
standard addition is used to achieve the best accuracy. Calibration by standard addition is
discussed in Chapter 20.
Introduction of the sample solution and subsequent atomization in a flame is a
complicated process made up of several steps as shown in Figure 9.3. In the first step,
the sample solution is dispersed as a fine aerosol of small droplets. The formation of small
droplets is termed nebulization. The aerosol is mixed with gaseous fuel and oxidant that
carry it into the flame. The solvent evaporates in the base region of the flame and the
resulting finely divided solid particles are carried to the center of the flame, which is the
hottest part of the flame. Here, the solid particles evaporate and as a result of atomization
the atomic vapor and elementary ions are formed. Finally, the atoms are carried to the
outer edge of the flame where oxidation may occur before the atomization products
disperse into the atmosphere. Most atoms are in their ground state and may absorb
electromagnetic radiation of a specific wavelength transmitted through the flame. Because
the velocity of the fuel/oxidant mixture through the flame is high, only a fraction of the
sample undergoes all these processes.
The atomization process is a consequence of the high temperature in the flame. The most
common fuel/oxidizer combination is acetylene/air, which produces a flame temperature of
2100–2400 K. This temperature is satisfactory for atomization of most elements and when a
hotter flame is required a combination of acetylene/nitrous oxide is used. With this
combination a flame temperature of 2600–2800 K is reached.
Nebulization
Evaporation of solvent
Evaporation of particles
Atomization
Wanted
Excitation
Unwanted
Ionization
Figure 9.3 General principle of atomization in a flame
118
Introduction to Pharmaceutical Chemical Analysis
9.2 Instrumentation
The outline of an atomic absorption spectrometer is shown in Figure 9.4. The radiation
source emits radiation of a wavelength that is specific for the element being analyzed and is
typically a hollow-cathode lamp. The hollow-cathode lamp contains a vapor of the same
element as that being analyzed and emits radiation of the same frequency as that absorbed by
the analyte element in the flame.
Unlike ordinary spectrometers, a background correction is necessary in atomic absorption spectrometers to distinguish signals generated by the hollow-cathode lamp and signals
generated by the flame. For this purpose beam chopping is most common. Segments of a
metal chopper are removed so that radiation passes through the device for half the time and is
reflected for the other half. Rotating the chopper at a constant speed causes the beam
reaching the flame to vary periodically from zero intensity to maximum intensity. As shown
in Figure 9.5 the beam of light from the lamp is periodically blocked by the rotating chopper.
Signals reaching the detector when the beam is blocked must be emissions from the flame.
Signals reaching the detector when the beam is not blocked must be from the lamp and the
flame. The difference between these two signals is the desired analytical signal, which is the
basis for the measurement of absorbance.
The samples, which are usually acidic aqueous solutions, are drawn into the flame
where the sample evaporates and is broken into free atoms. The beam of light from the
hollow-cathode lamp is directed through the flame. The flame replaces the cuvette in
conventional spectrometry and the path length of the flame is typically 10 cm.
A monochromator is used to select one line from the hollow-cathode lamp and to
reject as much emission from the flame as possible. The hollow-cathode lamp emits all
radiation lines for a particular element and among these one line has to be selected for
measurement of absorbance. The detector records the intensity of the radiation line when
a sample is drawn through the flame (I) and the intensity of the radiation line when no
sample is drawn through the flame (I0). The signals are converted electronically to the
corresponding absorbance units (log I0/I).
Absorption of radiation by the atomic vapor take the form of a series of narrow lines that
are a result of transition of an electron from the ground state to one or more higher energy
levels. One of these lines is selected for measurement. The wavelength of an absorbed line is
identical to the wavelength that is emitted when the electron returns from the excited state to
the ground state and the wavelengths lines are unique for each element. To measure
absorption with a high degree and sensitivity and specificity the radiation source should emit
Hollow-cathode
lamp
Chopper
Monochromator
Flame
Figure 9.4 Outline of a single-beam atomic absorption spectrometer
Detector
Atomic Spectrometry
Lamp
Chopper
119
Flame
Radiation from
the flame
Radiation from the
flame and the lamp
Figure 9.5 Modulating the beam from the hollow-cathode lamp
radiation of the same wavelength lines that are able to excite the atoms. The most useful
radiation source is the hollow-cathode lamp as shown in Figure 9.6.
The lamp is filled with an inert gas such as argon and consists of a cathode and an anode.
The cathode is usually fabricated from the analyte element. When a high voltage is applied
between the anode and the cathode, the inert gas is ionized and positive ions are
accelerated towards the negatively charged cathode. When positive ions strike the
cathode, metal ions are released into the gas phase. These are excited by collision
with high-energy electrons and then emit radiation of the same wavelength lines as that
absorbed by the analyte atoms in the flame.
Anode
Window
Cathode
Inert gas
Figure 9.6 A hollow-cathode lamp
120
Introduction to Pharmaceutical Chemical Analysis
Flame
Burner head
Fuel
Oxidant
Flow spoilers
Capillary
tube
Drain
Nebulizer gas
Sample
solution
Figure 9.7 Schematic view of an atomization device
The hollow-cathode lamp therefore emits radiation at wavelengths that exactly match the
wavelengths that are able to excite the free atoms of that element. Consequently, different
lamps are required for each element. For example for the analysis of zinc (Zn), a Zn-coated
cathode is used and free Zn atoms in the flame can easily absorb the emitted radiation line at
a wavelength of 214.0 nm. A disadvantage of the hollow-cathode lamp is that only one
element can be analyzed at a time, although some lamps are made with more than one
element in the cathode.
The next component of the spectrometer is the atomization device as shown in Figure 9.7.
The sample solution is drawn into the capillary tube of the nebulizer by a rapid flow of
nebulizer gas (air or N2O) past the tip of the capillary. The liquid sample is broken into an
aerosol of fine droplets as it leaves the tip of the nebulizer. It is mixed with fuel (acetylene)
before the aerosol is directed to the flame. The aerosol consists of droplets of different sizes
and flow spoilers are used to block large droplets of liquid from entering the flame.
The liquid from large droplets is collected at the bottom of the spray chamber and flows out
to a drain container. Only the very fine droplets of the aerosol, oxidant and fuel are fed into
the burner. The burner provides a flame that is usually 5 or 10 cm in length, which is the path
length of radiation. The flame replaces the cuvette in a conventional spectrometry.
According to Beer’s law, absorbance is proportional to the path length and long path
lengths provide increased sensitivity.
The intensity of radiation directed to the flame (I0), is reduced in the flame because free
atoms absorb radiation. The intensity of transmitted radiation (I) is measured with a detector
and the signals are converted to absorbance unites [A ¼ log(I0 /I)] shown on the instrumental
display. As the flame itself emits radiation at a number of wavelength lines a monochromator
is necessary to select the wavelength of radiation directed to the detector. Most instruments
are equipped with an ultraviolet/visible grating monochromator as describe in Chapter 7.
Instruments for AAS are single-beam or double-beam instruments. Figure 9.4 shows a
typical setup for a single-beam instrument. Radiation is directed to the flame and the beam of
radiation is modulated with a beam chopper. The advantage of single-beam instruments is
their simplicity and since all of the radiation is directed to the flame, single-beam
instruments provide good sensitivity. A disadvantage of single-beam instruments is that
Atomic Spectrometry
Mirror
121
Mirror
Detector
Hollow-cathode
lamp
Chopper
Monochromator
Flame
Mirror
Figure 9.8 Double-beam instrument
these instruments cannot correct for variations in lamp intensity and variations due to the
detector and the electronic system. These variations may affect the absorbance readings.
The double-beam principle as shown in Figure 9.8 effectively compensates for
instrumental variations. The radiation from the hollow-cathode lamp is chopped and
mechanically split into two separate beams; one passes through the flame and the other is
directed around the flame. The latter acts as a reference beam providing a signal for the
intensity of the lamp. The absorbance can be determined as the ratio between the intensity
of the two beams and possible variations in absorbance readings due to instrumental
variations are compensated.
9.3 Applications and Performance
The principal use of AAS is in limit testing for metals in pharmaceutical ingredients prior to
their incorporation into formulations and in limit testing for metals in finished products. It is
also used for the quantitative determination of elements in a number of preparations. Unlike
methods for molecular spectroscopy (e.g., UV, fluorescence), AAS is not used as a detector
in chromatographic analysis.
In AAS with flame atomization the precision is usually in the range of 1% and the method
accuracy is also very good. AAS is a very sensitive technique and can be used to determine
elements even at very low concentration levels. It is shown in Table 9.1 that AAS with
flame atomization can determine common elements quantitatively down to 0.1–100 ng/ml
(ppb level).
The efficiency and reproducibility of the atomization step largely determines the
sensitivity, precision and accuracy of the method. In addition to atomization in a flame,
atomization can also be carried out in electrically heated furnaces. Graphite furnaces are the
most common. Electrothermal atomization in a graphite furnace atomizes the entire sample
and retains the atomic vapor in the light path for an extended period. Electrically heated
furnaces offers greater sensitivity than that afforded by flames and requires smaller sample
122
Introduction to Pharmaceutical Chemical Analysis
Table 9.1 Detection limits for common elements
Element
Ag
Al
Ba
Ca
Cd
Cr
Cu
Fe
Detection limit (ng/ml)
Element
Detection limit (ng/ml)
3
30
10
1
1
3
2
5
Hg
Mg
Na
Ni
Pb
Sn
Zn
500
0.1
2
5
10
20
2
sizes. Compared to flames, electrically heated furnaces are more costly and require more
operator skill to determine proper conditions.
9.4 Practical Work and Method Development
The quantitative determination of elements with AAS measures the absorbance of free
atoms in the flame and the absorbance is related to the initial concentration of elements
in the sample solution. The quantitative determination is based on a calibration curve
where absorbance is plotted as a function of known concentrations of reference
solutions. It is assumed that the atomization processes are the same for test solutions
and reference solutions:
1. Test solutions and reference solutions are drawn into the nebulizer and carried into the
flame at the same speed.
2. Atomization of the analyte element is unaffected by the analytical matrix.
When this is not the case, significant error may occur. Problems of type 1, when the sample
solution is drawn into the nebulizer, occur when the viscosity of test solutions and reference
solutions vary. Solutions of high viscosity are drawn up and nebulized more slowly
than solutions of lower viscosity. In those cases quantification by standard addition is
recommended. All measurements are then made on solutions with the same viscosity.
Problems of type 2, where atomization of elements is affected by the matrix
components, are often referred to as chemical interferences. Chemical interference is
caused by any component of the sample that decreases the atomization of the element to
be determined. Thermally stable products that are difficult to atomize can be formed, for
example, sulfate and phosphate reduces the atomization of calcium by forming nonvolatile salts. In those cases quantitative measurements are too low. Using flames with a
higher temperature can often eliminate chemical interferences. A higher temperature
may supply more energy and increase the efficiency of the atomization process.
Replacing acetylene/air with acetylene/N2O can increase the flame temperature. Releasing agents added to the test solution can also eliminate chemical interferences. For
example lanthanium (III) can be added to the sample solution to protect calcium from the
interfering effects of sulfate. Lanthanium (III) forms more stable compounds with
sulfate than calcium does.
Atomic Spectrometry
123
Monochromator
Detector
Flame
Figure 9.9 Principal view of an atomic emission spectrometer
9.5 Atomic Emission Spectrometry
Atomic emission spectrometry (AES) is a method that uses the emission of light to determine
elements in a sample. AES can determine the identity of an element, but the main
application area is the quantitative determination of elements. In AES the sample is
atomized and free atoms of the atomic vapor are transferred to the excited state. The
excited atoms are unstable and return to the ground state by emission of electromagnetic
radiation. Each element emits radiation at a characteristic wavelength and the intensity of
the emitted radiation is proportional to the amount of element in the sample. As shown in
Figure 9.9 the wavelength of the radiation is selected by a monochromator and a detector
records the intensity. A calibration curve is necessary to establish the relationship between
the intensity of the signal and the concentration of element.
In AES the sample is brought into a flame or plasma as a gas or as an aerosol. The heat
evaporates the solvent and breaks chemical bonds to create an atomic vapor. The temperature is high enough to cause not only dissociation into atoms, but also a significant amount of
collisional excitation and ionization of the sample atoms. As shown in Table 9.2 the number
of free atoms excited increases sharply with increasing temperature. AES with atomization/
excitation in a flame is called flame emission spectroscopy or flame photometry and can be
used to determine elements that are easy to excite, such as lithium, sodium and potassium.
Because of its convenience, speed and relative freedom from interferences, flame emission
spectroscopy has become the method of choice for these elements.
Most other elements need a higher temperature than that created in a flame to be
transferred to the excited state. A temperature of 4000–8000 K can be reached in plasma of
argon or helium. Plasma is an electrically neutral, highly ionized gas that consists of ions,
electrons and atoms. The energy that maintains the analytical plasma is derived from
Table 9.2 The relationship between temperature and excitation of free atoms
Element
Wavelength (nm)
Atoms excited (%)
2000 K
Sodium
Calcium
Zinc
589.0
422.7
213.9
3
1 10
1 105
7 1013
3000 K
4000 K
0.06
4 103
6 108
0.44
0.06
1 105
124
Introduction to Pharmaceutical Chemical Analysis
electromagnetic energy. The technique is called plasma emission spectroscopy. Plasma
emission spectroscopy can be used to determine most elements in the periodic system. The
detection limits for most elements are in the pg/ml to ng/ml range which are significantly
lower than those obtained with flame emission spectroscopy. By changing the wavelength
settings of the monochromator, the emission from several elements can be measured. This is
an obvious advantage as the content of several elements can be determined simultaneously.
There is a linear relationship between the intensity of emitted radiation and concentration of
element in a sample. An unknown concentration of an element can be read from a calibration
curve based on direct calibration or from a calibration curve based on standard addition.
Acidified water is the solvent of choice for preparing test and reference solutions because of
the high solubility of elements.
9.6 Instrumentation
Instruments for atomic emission spectrometry are similar in design to atomic absorption
spectrometers, except that in the former the flame acts as a radiation source. The hollowcathode lamp and the chopper are therefore unnecessary in instruments for AES.
The instruments consist essentially of a sample introduction and nebulization system,
an atomizer (flame or plasma) to generate and excite atoms, a monochromator, a detector
and a data acquisition unit.
Instruments for flame emission spectroscopy, flame photometers, are relatively simple in
design. The flame is similar to the flame in AAS. Air is used as oxidizer and common fuel
gases are hydrogen, acetylene, propane or butane. Because the temperature of a flame is low
(<3000 K), flames are not powerful enough to cause emission for many elements, and
consequently few elements can be determined. The atomic spectra emitted from flames have
the advantage of being simpler than those from other sources. Interferences from other
elements are less probable and wavelength selection can be made with simple filters.
The advantage is that the instruments are relatively cheap and are easy to use.
In addition to flame emission, spectroscopy instruments are also based on plasma
emission. The instruments are called inductively coupled plasma (ICP) atomic emission
spectrometers (AES). These instruments are significantly more complex than flame
photometers. In an argon plasma the temperature can reach 10 000 K and all the elements
in a sample emit radiation. The spectra from plasma are more complex than the spectra from
flames. To avoid interferences from irrelevant elements the use of a monochromator for
wavelength selection is essential. Inductively coupled plasma sources yield lower detection
limits than other emission sources. Compared to flame photometers the instruments for
ICP-AES are complex and expensive.
9.7 Summary
AAS is an important method in pharmaceutical analysis and is used for the quantitative
determination of elements in pharmaceutical ingredients and formulations. The test solution
is drawn into a flame and the components of the solution are atomized in the flame.
Atomic Spectrometry
125
Electromagnetic radiation with a wavelength line specific for the element to be determined
is directed through the flame and the absorbance is measured. Based on Beer’s law, the
measured absorbencies are converted to concentrations.
In atomic emission spectrometry the sample solution is drawn into a flame or plasma, and
the sample components are atomized. Emission from free atoms is measured and the
intensity is proportional to the concentration of the element in the sample. The technique is
used for the quantitative determination of elements in pharmaceutical ingredients and
finished products. In addition flame photometry is an important method in clinical chemistry
for the determination of elements, such as lithium, in blood samples from patients.
10
Chromatography
Chromatography is the name give to a particular family of separation methods where the
separation is based on differences in rates of migration when the sample components are
transported by a mobile phase through a stationary phase. The mobile phase can be a gas, a
liquid or a supercritical fluid and the stationary phase may be a solid, a liquid or a gel.
Chromatography is considered the most important analytical technique in pharmaceutical
analysis. This chapter looks at the theory of chromatographic separation and at important
parameters used to characterize a separation.
10.1 General Principles
In chromatography we have two phases: a phase that moves (mobile phase) and a phase that
is stationary (stationary phase). In Figure 10.1a the stationary phase is solid particles that are
packed into a tube, so that they remain stationary inside the tube. The tube with the
stationary phase is called a column. The mobile phase is pumped through the column at a
certain speed. In Figure 10.1b the mobile phase brings the components of a sample (A, B, C)
into the column where they come in contact with the stationary phase. If the stationary phase
retards the transport of the components through the column, they will migrate through the
column at a lower speed than the mobile phase itself. If the stationary phase retards the
components of the sample differently, the components will migrate with different speeds
through the column.
This is illustrated in Figure 10.1c. In this example component A is not retained by the
stationary phase and consequently A migrates through the column with the same speed as
the mobile phase. The stationary phase retards substance B so that its migration speed is
slowed down, and the drug substance B takes longer to migrate through the column than
substance A. Substance C is even more retarded than B and migrates at the slowest speed
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
128
Introduction to Pharmaceutical Chemical Analysis
(a)
Mobile phase is
pumped through the
column
(b)
A, B and C is brought
into the column
(c)
A, B and C is transported
through the column by
the mobile phase
A, B and C
C
B
A
Figure 10.1 Chromatographic separation of components A, B, and C
through the column. When the differences in retention are sufficiently large, the components
elute from the column at different times and are separated.
Differential migration of individual components through the column depends on the
equilibrium distribution of each component between the stationary phase and the mobile
phase. Therefore, differential migration is determined by those experimental variables that
affect this distribution, such as the composition of the mobile phase, the composition of the
stationary phase and the temperature. If we want to alter migration to improve separation,
we must change one of these variables. The equilibrium distribution is given by the
distribution constant:
KC ¼
Cs
Cm
ð10:1Þ
where Cs is the concentration of component in the stationary phase and Cm is the
concentration of component in the mobile phase. The distribution constant is also called
the equilibrium distribution coefficient. The speed by which each component migrates
through the column is determined by the number of molecules of that component which is in
the mobile phase at any instant, since sample molecules do not move through the column
while they are in the stationary phase. Retention of a component is therefore determined by
its distribution constant. Components with a large distribution constant have a large portion
of its molecules in the stationary phase and these components are strongly retained in the
column. Components with a small distribution constant have a small portion of its molecules
in the stationary phase and these are less retained.
A component appears to migrate smoothly through the column but at the molecular level
the migration of a compound is highly discontinuous. Each time a molecule is sorbed to the
stationary phase, the migration is temporarily stopped while other molecules of the same
kind pass over. Some of these are immobilized a moment later and are overtaken by the first
molecule and so on. Each molecule thus follows a rapid, random alternation between the
stationary and the mobile phase where the concentration of a compound in the stationary
phase and in the mobile phase is determined by the distribution constant.
In a chromatographic system the sample solution is introduced into the flow of mobile
phase using an injector, which is placed in front of the column. A detector detects
Chromatography
Injector
Column
C
129
Detector
B
A
Mobile
phase
Figure 10.2 Chromatographic system consisting of an injector, a column, and a detector
compounds that are eluted from the column. Such a configuration is shown in
Figure 10.2.
In a setup as shown in Figure 10.2 with an injector, a column and a detector, it is possible to
measure the detector response of a component eluting from the column as a function of time.
If the detector response to compounds eluted from the column is recorded as a function of
time after injection, the visual output of the separation is called a chromatogram.
The idealized chromatogram is a sequence of Gaussian peaks on a baseline.
Figure 10.3 shows a chromatogram of the sample containing components A, B and C.
The non-retained substance A that migrated through the column with the same speed as the
mobile phase is the first out. Substance C is the most strongly retained component and is
eluted last. Ideally the substances appear in the chromatogram as Gaussian peaks. The time
from injection of the sample to maximum detector response of a substance is called the
retention time, tR. The retention time for substance C is shown in Figure 10.3. In a given
chromatographic system the retention time of a substance is characteristic for its physicochemical and physical properties and is used for identification.
Although the substances of a sample solution are eluted from the column as symmetrical
Gaussian peaks with different retention times, it is not certain that they are separated
completely. For the substances to be separated completely the detector response must reach
a baseline between the chromatographic peaks. The two peaks shown in Figure 10.4a are not
completely separated from each other because the second peak starts rising before the first
peak reaches the baseline. The two peaks shown in Figure 10.4b are completely separated.
The retention times are the same as in Figure 10.4a, but the peaks are narrower. This
A
B
Time (min)
C
tR
Retention time
of component C
Figure 10.3 Chromatogram of the sample containing A, B, and C
130
Introduction to Pharmaceutical Chemical Analysis
Overlapping
peaks
(a)
(b)
Figure 10.4 Chromatogram of (a) two overlapping peaks and (b) two completely separated
peaks
illustrates that both sufficiently different retention times and sufficiently narrow peaks are
necessary for a complete separation. If peak widths are narrow, less difference in retentions
is needed.
Selectivity is another parameter that affects the separation of adjacent peaks. Selectivity
expresses the retention of a component relative to another closely eluting component as
(a)
(b)
Figure 10.5 Chromatogram of (a) first separation and (b) after altering selectivity
Chromatography
131
illustrated by the two chromatograms in Figure 10.5. The two peaks are overlapping in
Figure 10.5a and completely separated in Figure 10.5b because the retention of the second
peak relative to the first peak is larger. The following sections review common terms used to
characterize retention, peak width and selectivity.
10.2 Retention
A chromatogram of two components is shown in Figure 10.6. The first component is not
retained by the stationary phase. It migrates through the column with the same speed as the
mobile phase and is eluted at time tM which is the time taken to pass through the void volume
of the column, VM. VM is also called the hold-up volume and tM the hold-up time. The second
component is retained by the stationary phase and is eluted at retention time tR, in a
corresponding retention volume of mobile phase, VR.
The retention volume VR and the volume of the mobile phase required for elution of an
unretained compound VM may be calculated from the following equations:
VR ¼ t R F
ð10:2Þ
VM ¼ t M F
ð10:3Þ
and:
F is the volumetric flow rate of mobile phase passing through the column. When the flow rate
is 1 ml/min, tR is 6 min and tM is 2 min, the retention volume VR and the column void volume
VM may be calculated:
VR ¼ 6 min 1 ml=min ¼ 6 ml
ð10:4Þ
VM ¼ 2 min 1 ml=min ¼ 2 ml
ð10:5Þ
In a chromatographic system with a given stationary and mobile phase, the hold-up volume
and the retention volume are constant and independent of the flow rate of the mobile phase
tR
VR
tM
VM
Figure 10.6 Determination of retention time and retention volume
132
Introduction to Pharmaceutical Chemical Analysis
while the hold-up time and the retention time are inversely proportional to the flow rate of
the mobile phase. Increasing the flow rate from 1 ml/min to 2 ml/min results in a 50%
reduction of the retention times and hold-up times.
It has earlier been explained that retention of a compound is determined by its distribution
constant. The relationship between the retention volume and the distribution constant Kc is
given by:
VR ¼ VM þ KC VS
ð10:6Þ
This equation shows that the retention volume of a compound is equal to the hold-up volume
VM plus the additional volume of mobile phase required to elute the compound (KCVS). An
unretained compound has a KC value of 0. Large values of KC mean high affinity for the
stationary phase and strong retention of the compound.
To provide a measure of the substance’s retention it is common to use retention times. The
retention times are affected, however, by several factors such as the speed of the mobile
phase and the column length. A parameter that is independent of these factors are the
retention factor k. The retention factor k is defined as:
k ¼ amount of compound in stationary phase=amount of compound in mobile phase
and can be expressed as:
k ¼ Kc
Vs
Vm
ð10:7Þ
The retention factor shows how the amount of the drug substance is distributed between the
stationary phase and the mobile phase. The distribution depends on the distribution constant
and on the volumes of the stationary and the mobile phase. When the retention factor is equal
to 1, the compound will be distributed equally between the stationary phase and the mobile
phase. When the retention factor is 5, there will be five times more of the compound in the
stationary phase than in the mobile phase.
The retention factor of a component may be determined from the chromatogram using the
following equation:
k¼
tR tM
tM
ð10:8Þ
The expression tR tM is the time the compound stays in the stationary phase, while tM is
the time the compound stays in the mobile phase.
Assume that a compound has a retention time tR of 6 min and that tM is 1 min. The
retention factor k is (6 1)/1 ¼ 5. Of the total time the molecules spend in the column
(6 min) they will move with the mobile phase for one-sixth of the time and be temporally
stagnant on the stationary phase for five-sixths of the time.
Figure 10.7 shows a chromatogram of six compounds in which both the retention time and
retention factor are shown. The substances have retention times between 3.2 and 7.0 min and
retention factors between 1.3 and 3.8.
Chromatography
133
tR
tM
0
2
4
6
8
Time (min)
0
1
2
3
4
k
Figure 10.7 Chromatogram showing retention times and retention factors
10.3 Column Efficiency
The peak width expresses how efficiently a compound is transported through the column. If
the molecules to a small extent are spread throughout the column, peaks widths are small.
The narrower the peaks are in relation to their retention times, the more effective is the
column. Peak broadening throughout the column is expressed by the parameter N, called the
number of theoretical plates or plate number.
A Gaussian peak as shown in Figure 10.8 is characterized by its retention time and the
standard deviation s.
The plate number is defined as:
t 2
R
N¼
ð10:9Þ
s
In Gaussian peaks the peak width is 2s at the points of inflection, which are at 0.607 h. In
practice it is difficult to measure peak widths accurately at 0.607 h, and in the European
Pharmacopeia the peak width Wh at half-height, 0.5 h, is measured instead. The relationship
between Wh and s is given by:
Wh ¼ 2:354s
and N is estimated by the following equation:
2
tR
N ¼ 5:54
Wh
ð10:10Þ
ð10:11Þ
134
Introduction to Pharmaceutical Chemical Analysis
tR
1
2σ
Wh
h
0.607h
0.5h
W
Retention time
Figure 10.8 Parameters used to measure peak widths
Another way to estimate N is to measure peak width at the baseline, W, defined as the
width between tangents drawn through the inflection points as shown in Figure 10.8. The
relationship between W and s is:
W ¼ 4s
and the plate number N is calculated by the equation:
t 2
R
N ¼ 16
W
ð10:12Þ
ð10:13Þ
Assume that a peak has a retention time of 6.0 min and that the peak width at half peak height
is 0.5 min. The plate number N is then:
2
6
¼ 798
ð10:14Þ
N ¼ 5:54
0:5
If the peak width at half peak height is 0.2 min, N is:
2
6
N ¼ 5:54
¼ 4986
0:2
ð10:15Þ
The narrower peaks are in relation to the retention time the greater N will be. The greater N
is, the more efficient is the column to give narrow peaks in relation to the retention time.
Chromatography
135
(a)
(b)
Retention time
Figure 10.9 Chromatogram of a peak with a short retention time (a) and with a long retention
time (b)
A component has about the same plate number regardless its retention. On the chromatogram peaks with short retention times are narrow, and the peak widths increase as the
retention times increase. This is shown in Figure 10.9 as a chromatogram of a peak with a
short retention time (a) and with a longer retention time (b).
The quantity N is proportional to column length L, so that when other factors are equal, an
increase in L results in an increase in N. A doubling of column length results in doubling of
plate number. The proportionality of N and L can be expressed as:
H ¼ L=N
ð10:16Þ
where H is the height of a single theoretical plate. The quantity H is a measure of the
efficiency of a given column per unit length of the column. If the column above is 5 cm
(50 mm) and N is 798, H is:
H ¼ 50 mm=798 ¼ 0:06 mm
ð10:17Þ
Small values of H mean large N values and efficient columns. By using H instead of N, one
can compare the efficiency of columns with different length.
10.4 Selectivity
The relative retention of two adjacent peaks in the chromatogram is described by the
separation factor given by the equation:
¼ k2 =k1
ð10:18Þ
136
Introduction to Pharmaceutical Chemical Analysis
k1
k2
Retention time
Figure 10.10 Retention factors of two closely eluting peaks
where k2 is the retention factor of the later of the two eluting peaks, and k1 is the retention
factor of the first eluting peak as shown in Figure 10.10.
The separation factor is a measure of the selectivity of a chromatographic system. It is a
constant for a given set of analytical conditions and is independent of the column type and
column length. The separation factor has values greater or equal to 1.0. When ¼ 1,
separation is not possible. The larger the value of , the easier a separation becomes. If the
retention factors of two adjacent peaks in the chromatogram are 3.5 and 4.5, the separation
factor is:
¼ 4:5=3:5 ¼ 1:3
ð10:19Þ
Figure 10.11 shows how a separation can be improved by altering values of adjacent
peaks.
10.5 Peak Symmetry
The symmetry factor (As) is used to signify peak symmetry. The symmetry factor is
calculated using the following equation (see Figure 10.12):
w0:05
ð10:20Þ
As ¼
2d
where: w0.05 is the width of the peak at one-twentieth of the peak height and d is the distance
between the perpendicular dropping from the peak maximum and the leading edge of the
peak at one-twentieth of the peak height.
An As value of 1.0 signifies symmetry. When As > 1.0, the peak is tailing. When As < 1.0,
the peak is fronting. The peak shown in Figure 10.12 is a tailing peak.
Chromatography
137
(a)
(b)
Retention time
Figure 10.11 Improvement of a separation by altering values of adjacent peak: (a) first
separation and (b) after altering values
d
w0.05
Figure 10.12 Parameters used to calculated peak symmetry
138
Introduction to Pharmaceutical Chemical Analysis
10.6 Resolution
The degree of separation between two peaks is defined as the resolution, Rs:
Rs ¼ 1:18
tR2 tR1
wh1 þ wh2
ð10:21Þ
Figure 10.13 shows the parameters needed for calculating Rs, where tR2 and tR1 are the
retention times of peaks 2 and 1 and wh2 and wh1 are the corresponding peak widths at half
peak height.
The European Pharmacopeia uses the equation above. When the peak width is measured
between tangents drawn through the inflection points, the equation is:
Rs ¼
2ðtR2 tR1 Þ
w1 þ w2
ð10:22Þ
A value of Rs ¼ 1.0 corresponds to a peak separation of 94% and baseline separation
corresponds to an Rs value of 1.5. If the retention times of two peaks are 6.2 and 7.0 min and
their peak widths at half peak height are 0.4 and 0.5 min, the resolution is:
Rs ¼
2ð7:0 6:2Þ
¼ 1:8
0:4 þ 0:5
The resolution is larger than 1.5 and the peaks are completely separated.
tR1
tR2
Wh
1
Wh
2
W1
W2
Retention time
Figure 10.13 Parameters needed for calculating Rs
ð10:23Þ
Chromatography
139
To control the separation of closely eluting peaks the following equation shows how the
resolution is related to the adjustable chromatographic variables of retention factor, plate
number and separation factor:
Rs ¼
1 pffiffiffiffi ð 1Þ k
N
4
ðk þ 1Þ
ð10:24Þ
where k and N are the average values for the two peaks. The retention affects the resolution
by k/(k þ 1), which approximates 1 as k increases. The optimum resolution for most
separations falls in the range k ¼ 3–10. The resolution increases with the square root of
the plate number, thus the plate number must increase fourfold to increase resolution by
a factor of 2. Resolution is strongly influenced by the separation factor and increases
with increasing values of . Figure 10.14 shows baseline separation of two peaks and
how this separation may be affected by poor column efficiency, low selectivity and
inadequate retention.
Completely separated peaks
Inadequte retention
Low selectivity
Poor column efficiency
Figure 10.14 Effect of inadequate retention, low selectivity, and poor column selectivity
140
Introduction to Pharmaceutical Chemical Analysis
10.7 Chromatographic Techniques
In pharmaceutical analysis the most frequently used chromatographic techniques are based
on a liquid or a gas as the mobile phase. When the mobile phase is a gas, the technique is
called gas chromatography (GC), and when the mobile phase is a liquid the technique is
called liquid chromatography (LC).
Gas chromatography is used for the analysis of volatile compounds. In GC the stationary
phase is typically a non-volatile liquid that is added as a coating on the inside of narrow
columns, called capillary columns. Compounds are vaporized in an injector and are brought
to the column by the mobile phase called carrier gas. Compounds are separated according to
their volatility and their solubility in the stationary phase.
In LC the stationary phase is typically porous particles with a large surface area and with
active groups on the surface. In thin layer chromatography (TLC) the particles are scattered
over a plate, while in high performance liquid chromatography (HPLC) the particles are
packed in columns. Components that are transported by the mobile phase come in contact
with active groups on the surface of the particles. Retention is caused by interactions
between the component and active groups on the surface of the particles. The most
frequently used separation principles in pharmaceutical analysis are normal phase chromatography and reversed-phase chromatography. In normal phase chromatography the
stationary phase is more polar than the mobile phase. The stationary phase consists of porous
particles with polar groups on the surface, and the mobile phase is an organic solvent.
Retention is by polar interactions. In reversed-phase chromatography the mobile phase is
more polar than the stationary phase. The stationary phase is porous particles with nonpolar
groups on the surface. Retention is by nonpolar interactions. The mobile phase is typically
an aqueous solution.
In the following chapters, we will look closely at the separation principles and the
instrumentation used for thin-layer chromatography, liquid chromatography and gas
chromatography.
10.8 Summary
A chromatogram is the visual output of a separation and is a graphical representation of
detector response versus time. Idealized chromatograms are represented as a sequence of
Gaussian peaks on a baseline. Retention of a compound is reported as retention time (tR) or
as retention factor (k). Column efficiency is expressed as the plate number (N) or as the
height of a theoretical plate (H), and the separation factor () expresses the separation
selectivity. Compounds can be separated if they are sufficiently retained, if the column has a
sufficient plate number and if separation selectivity is sufficient. The degree of separation
between two peaks is defined by the resolution (Rs).
11
Chromatographic Separation Principles
Chromatographic techniques involve two phases of which at least one is mobile and the
other most often is stationary. The two phases move relative to each other and in most
cases the mobile phase moves through a bed of a stationary phase. The stationary phase
may be a piece of paper (paper chromatography), or particles of a suitable material
placed as a thin layer on an inert support (thin-layer chromatography) or packed into a
tube or column (column chromatography). Column chromatography can be executed
using a liquid (liquid chromatography) or a gas (gas chromatography) as the mobile
phase. When a mixture of analytes is introduced into these systems a number of separation
mechanisms influence the partition of analytes between the two phases. This chapter gives
an introduction to important separation mechanisms and how different types of analytes
can be separated.
11.1 General Introduction
The basic principle in chromatography is the partition of analytes between two immiscible
phases. When analytes have a different partition ratio between the two phases they can
be separated. The distribution constant of a solute depends on the interaction with the
molecules in the two phases. In gas chromatography (GC) little interaction takes
place between the analytes and the mobile phase which is an inert gas and in which the
single molecules are far apart. Therefore retention and separation in GC is controlled by the
choice of stationary phase and the instrumental set-up (see Chapter 14). In liquid
chromatography (LC; covering thin-layer, TLC, and high-performance, HPLC) interactions
of the analyte with other molecules take place in the mobile as well as in the stationary
phase leaving several more parameters to be controlled. This chapter discusses the most
important separation principles and the separation mechanisms involved.
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
142
Introduction to Pharmaceutical Chemical Analysis
11.2 Normal Phase Chromatography
Normal phase chromatography is the most common separation principle in TLC but can also
be performed in HPLC mode. A polar stationary phase is used along with a more nonpolar
mobile phase. A typical choice could be the use of silica as the stationary phase and a
heptane–ethyl acetate or a heptane–propanol mixture as the mobile phase. When analytes
enter such a system the analytes interact with both the stationary and the mobile phase.
Analytes with no affinity to (no interaction with) the polar stationary silica are not retained
and travel with the speed of the mobile phase. Analytes having polar functional groups have
a higher affinity to the polar silica and show retention in the system. Adsorption onto the
stationary silica is a reversible interaction, and an increase in the polar component of the
mobile phase increases the competition for the adsorption sites on the surface, thus
weakening the interaction of the analyte with the silica. When the elution strength of
the mobile phase is increased, the retention of analytes decreases.
11.2.1
Silica
Silica, or silica gel, is the most important stationary phase in normal phase chromatography.
Silica has strong adsorption characteristics and is among other things used as a desiccant and
many substances can be adsorbed onto silica. Normal phase chromatography is also called
adsorption chromatography. Silica is a partially dehydrated form of colloidal polymerized
silicic acid. Silicic acid, H2SiO4, does not exist as the free monomer but is available in the
form of a sodium silicate solution. When the sodium silicate solution is acidified the
polymeric silica is formed. Silica used for chromatography undergoes an extensive
purification process to remove metal impurities and is then pulverized, dried and fractionated into appropriate particle sizes. Silica has typically a large surface area in the range
200–800 m2/g. The large surface area is due to the structure of the silica being a porous
material similar to a sponge. The pore volume of 1 g is more than 0.7 ml. In practice this
means that silica is a porous skeleton
that is manufactured with a well defined pore diameter
typically in the range 60–150 A (one angstrom is 10–10 m) when used for chromatography of
small molecules. When the mobile phase flows through the silica, it enters the entire volume
between the particles and the whole pore volume inside the particles. This provides a
tremendous network of contacts between the stationary and the mobile phase and analytes
are exposed to this surface area of stationary phase.
The silica is covered with silanol groups on the surface (–Si–OH) which are the adsorptive
groups. The silanol groups make the surface polar and behave as weak acids. Figure 11.1
shows a schematic drawing of silanol groups on the surface of silica.
Some silanol groups provide hydrogen bonding interactions with nearby silanol groups
while other groups are isolated. In some cases two silanol groups are attached to the same
silicium atom. The structure differences between silanol groups mean that they have
different activity. If a silanol group forms a hydrogen bond to a neighbouring group it will be
less active in adsorption processes than an isolated one. Performing chromatography in a
highly activated normal phase system where the silica is dry and the organic mobile is
without water can be difficult to control as the most active silanol groups readily adsorb
polar molecules. If a small amount of water is introduced into the system the water is
adsorbed and gradually masks the active sites. Thus it is a good idea to include small
Chromatographic Separation Principles
H
3
1
OH
O
O
Si
O
Si
O
O
Si
O
O
O
Si
Si
Si
O
Si
4 OH
2
OH
HO
O
O
Si
O
Si
O
O
OH
Si
O
143
O
Si
O
O
+
Me
O
O
Si
O
Si
Figure 11.1 Silanol groups: (1) free (isolated) silanol, (2) germinal silanols, (3) associated
silanols, (4) metal activated silanol on the surface of silica
amounts of water or other polar solvent into the mobile phase to bring them in equilibrium
with the stationary phase. The addition of an amine or an acid to the mobile phase has the
same effect.
Silanol groups are weakly acidic and the differences in their structure also result in
differences in acidity. The pKa value of pure silica is therefore an interval from 6 to 7.
Furthermore, the silica has an isoelectric point at about pH 2 because of increasing surface
protonation of the polymer at low pH.
11.2.2
Interactions
The interactions between silanol groups on silica and functional groups on the substances to
be separated are called polar interactions and are divided into four main groups:
1. Dispersion interactions
Dispersion gives weak interactions and occurs when the silanol groups induce temporary
dipoles in an analyte molecule. This may provide electrostatic attraction between
analytes and the adsorbent resulting in retardation. The interaction is strongest for
compounds with easily polarizable electrons as, for example, found in the aromatics.
2. Dipole–dipole interactions
Analytes having permanent dipoles may induce dipole interactions. The interactions are
stronger than dispersion interactions and are strongest for compounds with large dipole
moment. Examples of groups that provide permanent dipoles are: CN > NO2 > C¼O,
CHO > COOR > halogen > OH > COOH > –O– > NH2 Analytes having nitrile or nitro
groups participate in the strongest dipole interactions.
3. Hydrogen bonding interactions
Proton donors and proton acceptors provide hydrogen bonding interactions. The more
acidic the proton donor and the more alkaline the acceptor, the stronger is the interaction.
Carboxylic acids and phenols are examples of strong proton donors and amines are
examples of strong acceptors. Hydrogen bond interactions are stronger than the interactions mentioned above. Silica has both proton acceptor and donor properties.
4. Ionic interactions
Ionic interactions can occur between basic analytes and the acidic silanol groups. Such
interactions are undesirable because they often cause problems during separation. Ionic
interactions can be very strong and comparable to covalent bonds (see Section 3.2) but
144
Introduction to Pharmaceutical Chemical Analysis
can be avoided either by masking the most acidic silanol groups or by selecting mobile
phases that prevent the strong ionic interactions. Ion exchange interactions are discussed
further in Section 11.7.
11.2.3
Order of Elution
Substances are separated according to the strength of their interaction with silica. In normal
phase chromatography the following equation shows the dependence of the eluting strength
of the mobile phase, the properties of the analyte and the properties of the stationary phase
on the distribution constant, KD, of the analyte:
logKD ¼ logVa þ aðS0 AS e0 Þ
ð11:1Þ
Va is the surface volume of the adsorbend (the volume of a mono molecular layer of solvent
adsorbed to the surface), a is an expression for the surface energy of the adsorbend, S0 is
the adsorption energy for the analyte when dissolved in pentane, As is the area taken up on
the adsorbend by one molecule of the analyte and e0 is the solvent strength of the mobile
phase. An adsorbend with a large surface area and having many adsorptive groups on
the surface provides a stronger retention. In general S0 increases with the polarity of the
analyte. If an analyte has more than one functional group the adsorption energy for the
analyte is equal to the sum of the adsorption energies for each functional group. Adsorption
energies for a number of functional groups are given in Table 11.1.
An analyte with a small area and a high number of polar functional groups has a high
distribution constant and thus a strong retention. This can be illustrated with the two
compounds given in Figure 11.2, where compound B has a stronger retention than
compound A.
Ionic and hydrogen bonding interactions are the strongest interactions, while the weaker
dipole interactions are stronger than dispersion interactions. Saturated hydrocarbons do not
Table 11.1 Adsorption energies for analytes with a functional group attached to an aliphatic or
an aromatic moiety. The data applies for a system with silica as adsorbent and pentane as solvent
Functional group
Methyl
Aromatic carbon
Fluoro
Chloro
Thiol
Ether
Tertiary amine
Aldehyde
Nitro
Nitrile
Ester
Keto
Hydroxy
Primary amine
Carboxylic acid
–CH3
–CH¼
–F
–Cl
–SH
–O–
–N¼
–CHO
–NO2
–CN
–COO–
¼CO
–OH
–NH2
–COOH
Aliphatic
Aromatic
0.07
0.11
0.25
0.15
0.20
0.67
0.87
1.54
1.74
1.70
3.61
5.80
4.97
5.71
5.27
5.27
5.27
5.60
8.00
7.60
3.48
2.77
3.33
2.68
2.68
4.20
5.10
6.10
Chromatographic Separation Principles
COOH
COOH
A
OH
OH
B
F
F
145
Figure 11.2 Structures of compound A (diflunisal) and compound B (salicylic acid)
have any interactions with silica and thus have no retention. Retention increases as the
interaction with silica becomes stronger. Aromatic hydrocarbons only have poor retention
due the weak dispersion interactions. Interaction and retention increases with the polarity of
the functional groups. This is illustrated in Figure 11.3. Alcohols and phenols are more polar
than the compounds with carbonyl and are retarded stronger because of hydrogen bonding
interactions with the silanol groups. Primary amines and carboxylic acids provide the
strongest hydrogen bonding interactions and are retarded strongly by silica.
Retention increases with the number of functional groups. A substance with two hydroxyl
groups is retarded more than a drug with only one hydroxyl group.
11.2.4
Other Stationary Phases
Alumina and magnesium silicates are examples of other polar materials to be used as
stationary phases in normal phase LC. The silanol groups on silica can also be derivatized
with ligands containing other polar groups such as diol, CN and NH2. These polar materials
can provide changes in selectivity and thus changes in the order of elution compared to that
obtained on silica.
Interaction
with silica
Amines and carboxylic acid
Alcohols,
Esters, aldehydes
phenols
ketones
Ethers
Aromates
Saturated
hydrocarbons
Retention
Figure 11.3 Retention of different functional groups as a function of the interaction with silica
146
11.2.5
Introduction to Pharmaceutical Chemical Analysis
Mobile Phases
In normal phase chromatography the mobile phase is more nonpolar than the stationary
phase and therefore made of organic solvents. The intention of the mobile phase is to:
.
Keep the analytes in solution;
Transport the analytes through the bed of stationary phase;
. Contribute to the separation;
. Compete with the analytes for the adsorption sites on the stationary phase.
.
Silica is a polar adsorbent providing strong interactions and strong retention of polar
compounds and weak interactions and low retention of nonpolar substances. Mobile phases
violating strong interactions are called strong mobile phases and mobile phases that only
break weak interactions are called weak mobile phases.
The strength of the mobile phase is decided by the polarity of the solvent used, and
solvents can be ranked according to their solvent strength, e (see Table 11.2). Solvent
strength increases with increasing polarity. Increased solvent strength of mobile phase
decreases the retention of the substances.
As shown hydrocarbons have e close to 0 and they have the lowest solvent strength.
Hydrocarbons cannot break polar interactions and are therefore the weakest solvents.
Solvent strength and e value increases with increasing polarity of solvents. Methanol has e
at 0.73 and, together with water, is considered to be the strongest solvent.
In normal phase chromatography the mobile phase seldom consists of a single solvent.
Usually it consists of a mixture of two or three solvents. The reason is that it is easier to
control the solvent strength by mixing solvents than by trying out a number of single
Table 11.2 Solvent strength e (silica) for
selected solvents
Solvent
e
n-Pentane
n-Heptane
Cyclohexane
Toluene
Chloroform
Dichloromethane
Triethylamine
Diethylether
Ethyl acetate
Methyl t-butylether
Tetrahydrofuran
Acetone
Dimethylsulfoxide
Acetonitrile
2-Propanol
Diethylamine
Ethanol
Methanol
Acetic acid
Water
0.0
0.01
0.04
0.22
0.26
0.30
0.42
0.43
0.48
0.48
0.48
0.48
0.50
0.52
0.60
0.63
0.68
0.73
High
High
Chromatographic Separation Principles
147
Table 11.3 Solvent strength of the volume percentage of dichloromethane, ethyl acetate and 2propanol mixed with a hydrocarbon (e.g., n-heptane)
e
Dichloromethane
Ethylacetate
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.50
0.60
3.5
8
17
31
55
100
0.3
0.7
1
1.7
5
15
30
48
100
2-Propanol
0.15
0.3
0.5
1.5
4
14
35
100
solvents. By mixing a solvent with a high e value with a solvent with a low e value, a large
area of solvent strength can be covered. Examples of some solvent mixtures are shown in
Table 11.3. By mixing n-heptane with dichloromethane an e value in the range from 0 to 0.3
can be obtained, while mixtures of n-heptane and 2-propanol gives e ranging from 0.3 to 0.6.
Retention decreases with increasing solvent strength and a reasonable choice of e in the
mobile phase ensures that drug substances receive adequate retention. However, not all
ingredients of the sample are separated from each other even if the e value of the mobile
phase is satisfactory. In such a case, try other mobile phases, with the same solvent strength,
but with a different composition to obtain separations with sufficient selectivity.
Alkaline substances can cause unwanted ionic interactions with the silanol groups since
the silica has a pKa value between 6 and 7 (Figure 11.4).
Some of the silanol groups are negatively charged at neutral pH and participate in ionic
interactions with positively charged analytes. Ionic interactions are much stronger than
other polar interactions. It can therefore be difficult for solvents to displace the analytes from
the strong adsorption sites at the stationary phase, resulting in peak tailing of the analyte.
The problem can be overcome by adding an amine like diethylamine to the mobile phase.
The amine then competes with the analytes for the adsorption sites and displaces the analyte,
forming a more gaussion peak.
Another way to overcome peak tailing of basic analytes is to add a carboxylic acid to
the mobile phase. The chromatography is now performed at a lower pH which suppresses
the ionization of the silanol groups, reducing ionic interactions. Figure 11.4 gives an example
of this.
11.2.6
Summary of Normal Phase Chromatography
In normal phase chromatography the stationary phase is more polar than the mobile phase.
Organic solvents are used for the mobile phases and silica is the most common stationary
phase. Retention of analytes is determined by the properties of the compounds, the nature of
the stationary phase and the mobile phase. The most polar substances are the most strongly
retarded. Retention changes with the strength of the mobile phase. The highest retention is
obtained using nonpolar solvents with low solvent strengths. An increase in the polarity of
the solvents increases the solvent strength and the retention decreases.
148
Introduction to Pharmaceutical Chemical Analysis
+ 0.016 M
Propionic
acid
AU
No
acid
8
16
24
32 min.
Figure 11.4 Chromatogram of hydroxyatrazin with tailing and after addition of a carboxylic
acid. HPLC system: column silica 120 4.6 mm, 5 mm i.d., with dichloromethane þ methanol
(95 : 5, v/v) as mobile phase without and with propionic acid added
11.3 Reversed Phase Chromatography
Reversed phase chromatography is the most important separation principle in liquid
chromatography. In reversed phase chromatography the stationary phase is hydrophobic,
and the mobile phase is a more polar aqueous solution. This section focuses on different
stationary phases, mobile phases and the separation principle used in reversed phase
chromatography.
11.3.1
Stationary Phases
Column packing materials used as stationary phases for reversed phase chromatography are
typically made of silica derivatized with reagents to form a more or less hydrophobic surface
on the silica particle. They are typically obtained by binding hydrophobic groups to the
silanol groups using chlorosilanes or other organic silane reagents.
The examples of stationary phases shown in Figure 11.5 are ranked after declining
hydrophobicity. Octadecyl (C18) is the most hydrophobic phase and cyanopropyl (CN) is
the least hydrophobic of these phases. By far, most of the chromatography performed in
reversed phase mode makes use of C18 materials, which are also known as octadecyl or
ODS (octadecylsilane) materials.
The surface of silica is covered with silanol groups placed fairly close together. The
derivatizing silane reagent has three alkyl groups connected to the silicium atom and is
therefore a bulky group. Due to steric hindrance it is not possible to react all silanol groups
Chromatographic Separation Principles
149
R'
O
O
Si
O
R'
OH + Cl
Si
R
Si
O
R
R'
+
HCl
O
R'
O
Si
O
R' = Typically CH3
R = CH3-(CH2)17-; Octadecyl
CH3-(CH2)7- ; Octyl
C6H5-(CH2) 3-; Phenyl
CH3 - (CH2)3- ; Butyl
CN- (CH2)3- ; Cyanopropyl
Figure 11.5 Derivatization of silica with a chlorosilane reagent. Varying the substituents R 0 and
R produces stationary phases with different properties
with the reagent. A significant percentage of the silanol group can therefore still be present
after derivatization and in order to minimize the number of free silanol groups the material
can then be treated with trimethylchlorosilane. This process is named endcapping. However
even after endcapping some free silanol groups will still be present and accessible for polar
interactions (Figure 11.6), for example, ionic interactions with amines.
Silica-based stationary phases can be used with mobile phases with pH in the range 2–8.
Silica dissolves in the alkaline environment above pH 8 and also to some extent in an acidic
environment with pH < 2. It is the siloxane bonds that are cleaved. However, it is now
possible to obtain silica-based polymers stable in the whole pH range by including ethan
O O
Si
Si
O O
O Si OH
O
Si OH
O O
Si
O Si O
O
(a)
O O
Si
Si
O O
O Si OH
O
Si O Si
O O
Si
O Si O
O
(b)
Figure 11.6 C18 column packing material (a) before and (b) after endcapping with
trimethylchlorosilane
150
Introduction to Pharmaceutical Chemical Analysis
H
C
C
H
H2
C
H
C
H2
C
H
C
H2
C
H2
C
H
C
H2
C
H
C
H2
C
H
C
C
H
H2
C
H
C
H2
C
H2
C
H
C
H2
C
Figure 11.7 Polystyrene–divinylbenzene copolymer
ebridges in the silica polymer. These materials can be derivatized in the same manner as
described above. When selecting a C18 material for chromatography it is important to be
aware of the pH interval in which it can be used. When performing chromatography outside
the pH range 2–8 another possibility is to use pH stable organic polymers like polystyrene–divinylbenzene (PS-DVB). The structure of polystyrene–divinylbenzene copolymer is
shown in Figure 11.7.
Polystyrene–divinylbenzene can be used in the pH range 1–13 and it often provides a
stronger retention of analytes compared to silica-based C18 materials, but the column
efficiency of the organic polymeric stationary phases is less than for silica-based materials.
Furthermore, the organic polymeric stationary phases are less rigid than silica and only
highly crosslinked materials are suitable for HPLC.
Activated carbon is an example of another nonpolar adsorbent. Hydrophobic substances
in an aqueous environment adsorb onto the hydrophobic surface of the carbon. Recently also
diamond, which is also made of carbon, has been introduced as a column packing material
for reversed phase chromatography. This latter material is however not often used in
pharmaceutical analysis.
11.3.2
Retention Mechanisms
The main separation mechanism in reversed phase chromatography is hydrophobic
interaction. Nonpolar analytes are therefore retained strongly while more polar analytes
elute earlier. Octadecylsilane (C18) column packing materials are the most hydrophobic of
the usual commercial stationary phases available. The main forces of interaction are van der
Waal’s forces, which are relatively weak forces, but are present in a high number per
molecule, and interaction thus increases with molecular size. The interactions take place
between the hydrophobic hydrocarbon chains of the stationary phase and the hydrophobic
parts on the analyte molecule, as shown for naproxen in Figure 11.8.
Even highly polar analytes have a small retention due to the presence of minor
hydrophobic parts in the molecule. Organic analytes having the same functional group
are separated according to the size of the hydrophobic moiety, thus an additional CH2 group
Chromatographic Separation Principles
O O
Si
Si
O O
O Si OH
O
Si O Si
O O
Si
O Si O
O
151
CH3
COOH
H3 CO
Figure 11.8 Hydrophobic interaction between the hydrocarbon chain of C18 material and the
hydrophobic parts of naproxen
increases retention. This is illustrated in Figure 11.9. Analyte B is retarded more strongly
than analyte A because it has a longer hydrophobic side chain.
Besides the hydrophobic interactions with the hydrocarbon chains a secondary retention
mechanism can be displayed by the remaining silanol groups. Polar analytes especially
amines can have a high affinity for the silanol groups besides the hydrophobic interactions. It
is therefore of importance how many remaining silanol groups are present and how well they
are covered.
For substances with ionizable functional groups, retention is dependent on whether these
groups are ionized or not. Retention decreases with increasing ionization and this is of
(a)
(b)
NH2
NH2
H 3C
H 3C
Figure 11.9 Chromatogram of two analytes with a different size of side chain. The separation
was obtained using reversed phase chromatography
152
Introduction to Pharmaceutical Chemical Analysis
Table 11.4 Mobile phases with similar eluting strengths
Mobile phase
A
Methanol
Acetonitrile
Tetrahydrofuran
Water
B
C
60
46
40
54
37
63
course a function of the change in the mass distribution ratio for such analytes at different
pH values.
11.3.3
Mobile Phases
Mobile phases for reversed phase chromatography consist of mixtures of water and one or
more organic solvents that must be miscible with water. The organic solvents used are called
organic modifiers as they modify the strength of the mobile phase. Increased content of
organic modifier increases the strength of the mobile phase and retention of analytes
decreases. The solvent strength of methanol is somewhat weaker than that of acetonitrile
which is weaker than tetrahydrofuran. Mixtures A, B, and C (shown in Table 11.4) have
about the same solvent strength, they are isoeluotropic.
A mobile phase of 60% methanol in water have about the same eluting strength as 46%
acetonitrile in water or 37% tetrahydrofuran in water. When performing a separation of a
mixture of compounds using the isoeluotropic phases A–C, the average retention of the
compounds is about the same magnitude for the three phases. However, the order of elution
between the substances can be somewhat different because the selectivity in the three
systems can be different. Changing the organic modifier can thus be used to change the
separation selectivity (Figure 11.10).
Methanol is cheaper and less toxic than acetonitrile. The main drawback when using
methanol as modifier is the formation of more viscous mixtures with water creating an
increase in back pressure in the HPLC system.
Acetonitrile may be better suited for UV detection at low wavelength and it does not form
viscous mixtures with water (Figure 11.11).
Retention of neutral substances is only controlled by the content of organic modifier in the
mobile phase and is not affected by pH. For substances with ionizable functional groups
retention are dependent on both the content of organic modifier in the mobile phase and pH.
pH in the mobile phase is controlled by the addition of buffers. The buffer concentration in
the final mobile phase is typically in the range from 0.01 to 0.05 M. Phosphate buffers have
for many years been the first choice because phosphate buffers have good buffering
properties and low UV absorbance. A problem with phosphate buffers is that they have
poor solubility in organic solvents and thus can precipitate at high concentration of organic
modifier in the mobile phase. After the introduction of mass spectrometers as routine
detectors in LC there was a change towards volatile buffers of organic acids such as acetic
acid or formic acid and their ammonium salts. These buffer substances also have a better
solubility in organic solvents.
Chromatographic Separation Principles
50% methanol
2
153
35% acetonitrile 4+5
2
1
5
1
3
4
6
3
6
7
0
5
7
10
0
5
10 min
Figure 11.10 Example of solvent selectivity. Separation of seven test solutes on a C18 reversed
phase HPLC column using methanol or acetonitrile as organic modifier: (1) benzylalcohol,
(2) acetophenone, (3) phenylethanol, (4) propiophenone, (5) anisole, (6) toluene, (7) p-cresole
For the separation of acids or bases, changing the pH of the mobile phase can cause large
changes in separation selectivity. Figure 11.12 shows how the retention of acids and bases
varies with pH. Retention decreases with increasing ionization.
If an acid in the sample has a pKa value of 4.5 it is 50% ionized at pH 4.5 and fully ionized
at pH > 6.5. The partition between the mobile and stationary phases is shifted at high pH
towards the mobile phase, resulting in a low retention. Retention varies greatly in the range
around the pKa value, where small changes in pH provide major changes in ionization and
retention. The weak acid is retarded most strongly when ionization is suppressed at low pH.
The same consideration is true for a weak base at high pH. In this example, it has a pKa value
of 9.5 Retention is strongest when ionization is suppressed, at pH 9.5 half of the molecules
are ionized, and pH < 7.5 it is fully ionized. Many silica-based column packing materials
cannot be operated above pH 8.0 and therefore amines are often chromatographed at about
pH 3–4. Retention is then unaffected by pH as long as the base is fully ionized. In the pH
range where drugs are either fully ionized or where the ionization is completely suppressed,
the retention is not affected by minor changes in pH.
To obtain robust analytical methods, it is important to choose a pH where retention only
varies little with small changes in the composition of the mobile phase. It is unfavorable to
select a pH near the pKa value, unless it is a question of the separation of two closely related
substances with a small difference in pKa values. Typically, acids are separated by a pH that
suppresses ionization and bases at a pH where they are ionized. When using column packing
materials stable in the entire pH range it is of course possible to chromatograph bases in the
ion suppressed mode.
154
Introduction to Pharmaceutical Chemical Analysis
EtOH
2.5
THF
Viscosity (cp)
2.0
MeOH
Acetone
1.5
1.0
MeCN
0.5
0
10
20
30
40
50
60
70
Volume % water
80
90
100
Figure 11.11 Viscosity of mixtures of water and organic solvents. EtOH ¼ ethanol, THF ¼
tetrahydrofuran, MeOH ¼ methanol, MeCN ¼ acetonitrile
6
4
k
2
2
4
6
8
10
12
pH
Figure 11.12 Variation of the retention factor, k, for a weak acid (– – –) and a weak base (——)
as a function of pH in the mobile phase
Chromatographic Separation Principles
11.3.4
155
Ion-Pair Chromatography
In the previous section it is recommended to separate the substances either at a pH which
suppresses the ionization or at a pH where the substances are completely ionized. In the
latter case the retention may prove to be too small. In that case ion-pair chromatography
should be considered. The technique is particularly useful for the separation of basic
compounds in reversed phase chromatography.
In ion-pair chromatography the analytes have to be ionized and therefore pH has to be
controlled. The mobile phase typically has a pH of 7 where both carboxylic acids and
aliphatic amines are ionized. All ions in the system are surrounded by ions of opposite
charge in order to keep electro neutrality in the system. Buffer ions are normally very polar
and have only little interaction with the analytes. But when larger ions are added to the
mobile phase they can form ion pairs with other ions of opposite charge in the system. In ionpair chromatography larger hydrophobic ions are added to the mobile phase. The formed
hydrophobic ion pairs appears neutral to the surrounding. The ion-pairing process is a
dynamic process where molecules are exchanging all the time but when adding the ionpairing reagent, the counter ion, in excess, there is a high probability for the formation of an
ion pair with the analyte. The ion pair is hydrophobic and has a retention depending on the
nature and concentration of the counter ion.
When the molecular size or the concentration of the counter ion increases the
retention increases. However, the effect of increasing concentration of the counter ion
decreases at higher concentrations and therefore concentrations in the range 1–5 mM is
recommended.
Sulfonic acids and perfluorocarboxylic acids are common counter ions for protonated
bases, and quaternary ammonium compounds are commonly used for ionized acids.
Figure 11.13 shows the structure of three typical counter ions, octanesulfonate, heptafluorobutyrate and tetrabutylammonium. Octanesulfonic acid has a pKa below 1, is negatively
charged throughout the pH range used and forms ion pairs with positively charged bases.
Heptafluorobutyrate behaves similarly and can also be considered negatively charged
throughout the pH range. Tetrabutylammonium is a quaternary ammonium compound
that is positively charged in the full pH range. The ion is neutral similar to a sodium ion.
The positively charged group provides ion pairs with negatively charged acid groups.
Octanesulfonate provides larger retention of bases than pentanesulfonate, and tetrabutylammonium ions provide larger retention of acids than tetrametylammonium ions.
Retention in ion-pair chromatography is increased by changing the mobile phase as
follows:
.
Reduce the concentration of organic modifier;
Increase the concentration of the counter ion;
. Increase the molecular size of the counter ion.
.
11.3.5
Summary of Reversed Phase Chromatography
In reversed phase chromatography the stationary phase consists of hydrophobic groups
placed on the surface of a solid matrix. The solid matrix is most often silica onto which the
stationary phase is chemically bonded. The mobile phase is typically an aqueous buffer
solution mixed with an organic modifier such as methanol or acetonitrile. Retention of
156
Introduction to Pharmaceutical Chemical Analysis
H2
C
H 3C
H2
C
C
C
H2
H2
Octanesulfonate
F2
C
F3C
H2
C
SO2OC
H2
CH3
H2C
H2
C
H 3C
N
C
H2
COO-
C
F2
Heptafluorobutyrate
CH2
+
H2
C
C
H2
H2C
CH3
CH2
H3C
Tetrabutylammonium
Figure 11.13 Structures
tetrabutylammonium
of
the
ions
octanesulfonate,
heptafluorobutyrate,
and
neutral compounds is not affected by pH, while the retention of ionic compounds increases
when the ionization of the analytes decreases. Silica-based stationary phases can normally
be used in the pH range 2–8. If satisfactory separations are not achieved in this pH range,
organic polymers as stationary phases or other more pH-stable column packing materials
can be used. Alternatively, ion-pair chromatography may solve the problem.
11.4 Hydrophilic Interaction Chromatography
Hydrophilic interaction chromatography (HILIC) is a chromatographic principle used for
the separation of relatively polar analytes where the reversed phase mode does not provide
sufficient retention. The mobile phases used in HILIC often have a content of organic
solvent (e.g., acetonitrile) between 80 and 100%. The stationary phases are very polar [e.g.,
silica or silica derivatized with polar groups (diols, twitter ions, etc.)]. The systems can
therefore be considered as highly deactivated normal phase systems, where a hydrophilic
stationary phase is present. The less water present in the mobile phase the stronger is the
retention of the analytes. Thus, water is the strongest solvent in HILIC. If gradient elution is
applied in HILIC it is initiated using a high content of organic solvent and progresses with
increasing amounts of water.
The HILIC mode has become very popular for analysis of polar substances where
reversed phase chromatography has been less successful. The high content of organic
solvent eases the spray in electrospray mass spectrometry and thus improves sensitivity in
LC-MS methods, and although HILIC is a normal phase mode the content of water in the
mobile phase still makes it possible to use this mode for bioanalysis, which is more
problematic when using non-aqueous normal phase chromatography.
11.5 Chiral Separations
Section 3.3 gives a short discussion of stereochemistry. From this it is evident that it is a
challenge to separate pairs of enantiomers as they have the same physicochemical
Chromatographic Separation Principles
157
characteristics apart from their ability to rotate the plane-polarized light. If two substances
have the same distribution constants they cannot be separated in a chromatographic system.
Two enantiomers can therefore only be separated on chromatographic systems if it is
possible to introduce a difference in their distribution constants. This can in principle be
done in two ways: an indirect way and a direct way.
In the indirect way the enantiomers are derivatized with an optically active reagent to
form diastereomers. Diastereomers have more than one chiral center and have different
physicochemical characteristics and thus also different distribution constants. It is therefore
possible to separate enantiomers after derivatization to diastereomers in a standard
chromatographic system. This can either be by LC or GC.
When performing derivatization with chiral reagents it is important to be aware of
the purity of the derivatization reagent. The sample (S) being a mixture of the R- and
S-enantiomers of the analyte (e.g., 1% of one in the other) is derivatized with the optical
active reagent (R).
The ideal derivatization process with 100% pure reagent (R-R) results in two diastereomers:
Ideal derivatization process : R; S-S þ R-R ! R-S; R-R þ S-S; R-R
However, if (when) the reagent contains an enantiomeric impurity the reaction looks
like this:
Derivatization process : R; S-S þ R; S-R ! R-S; R-R þ S-S; S-R þ S-S; R-R þ R-S; S-R
The four different substances achieved are mirror images of each other (enantiomers) in
pairs and thus result in only two chromatographic peaks. If an analyte contains 1% of the
enantiomer and it is analyzed by chiral derivatization followed by achiral HPLC analysis
using a chiral derivatization reagent containing 1% of its enantiomer, the bias of the analysis
is about 100%.
Enantiomeric purity ð%Þ ¼
Area ð%Þ of major peak 100
enantiomeric purity of reagent ð%Þ
ð11:2Þ
where the area (%) of major peak is the area of the major peak as a percentage of the total
area of the two diastereomers in the chromatogram. It is therefore very important to use a
reagent of high chiral purity – preferably always above 99.5% purity.
Often the task is to separate a small amount of one enantiomer in the presence of the other
enantiomer. If the derivatization results in a given order of elution of the two peaks it is
possible to reverse this order of elution if the other chiral enantiomer of the reagent is used
for derivatization. In this way it is possible to elute the minor peak before the major peak and
in many cases this improves the quality of the quantitative measurement.
When using direct separation of enantiomers they are injected directly into the chromatographic system similar to what is used in other chromatographic methods. The direct
chiral separation (without derivatization) of enantiomers is only possible if chirality is
introduced into the chromatographic system. Chirality of chromatographic systems is
achived using chiral mobile or stationary phases. In GC it is only possible to have a chiral
stationary phase and this can be performed using either phases based on cyclodextrins or
phases based on polymers into which optically active amino acids have been incorporated.
In LC chirality can be introduced in the mobile or in the stationary phase and both modes
158
Introduction to Pharmaceutical Chemical Analysis
have been applied. If chirality is in the mobile phase diastereomeric complexes with
different complex constants have to be formed to achieve separation on an achiral stationary
phase. A chiral reagent is added to the mobile phase, and due to the different complex
constants of the diastereomeric complexes formed with the enantiomers to be separated,
they can be separated on an achiral stationary phase.
But more often chiral stationary phases have been used and a huge number are available
on the market. The phases consist of a polymer (silica, cellulose, methacrylate) to which a
chiral molecule is attached. The chiral molecules used are very different in nature (proteins,
polysaccharides, cyclodextrins, antibiotics, helical methacrylates, etc.) and therefore also
provide different selectivity. Unfortunately, the achievement of chiral separation is still
much of a trial and error and therefore it is important to consult literature and vendors of
stationary phases before starting the experiments.
11.6 Size Exclusion Chromatography
Size exclusion chromatography (SEC) is a separation method which separates substances
according to their molecular size. The principle of SEC is based on a sieving effect and
not on partition between immiscible phases. In this section we look briefly into the
separation principle.
11.6.1
Principle
In SEC the analytes are separated according to their molecular size in solution. It is very
important that the analytes are very soluble in the mobile phase used. In this separation
system the stationary phase is identical with the mobile phase, and the stationary phase is
held within the pores of a totally porous matrix. The liquid phase surrounding the matrix
particles is identical with the stationary phase. To obtain true SEC the analytes must not
interact with the matrix particles. If interaction between the analytes and the matrix particles
takes place it can be considered as a kind of adsorption chromatography. This compromises
the SEC analysis.
In order to separate the hydrophobic substances organic solvents are used as mobile
phases and the technique is called gel permeation chromatography (GPC). For watersoluble substances aqueous mobile phases are used and the technique is called gel filtration
chromatography.
Separation occurs when analytes penetrate into the pores of the matrix. The more of the
pore volume the analytes “see” the more they are retained. Analytes that can penetrate to the
full pore volume are eluted corresponding to the total volume of liquid in the column. This
volume corresponds to the hold-up volume in a standard LC system. This is illustrated in
Figure 11.14. Molecules not able to penetrate into the pores are eluted corresponding to the
volume of liquid surrounding the matrix particles.
The large molecules pass through the column outside of the particles, at the same speed as
the mobile phase. The volume of mobile phase that carries them through the column is called
the exclusion volume, VO. The exclusion volume is the volume of mobile phase between the
particles in the column. The smallest drug molecules are transported by the mobile phase
Chromatographic Separation Principles
LC
Detector response
VR
1
VR
159
2
VM
SEC
106
105
104
103
Mol. weight
107
102
V0
VM
VI
VR
Figure 11.14 LC versus SEC and calibration curve for SEC
into the smallest pores of the particles. Since the path length through all the pores of the
packing material is much longer, the small molecules get a longer retention time. The
volume of mobile phase that is used to elute these substances is called the total permeation
volume, VM. Figure 11.14 shows a chromatogram and a calibration curve.
The matrix particles for SEC are manufactured with controlled pore sizes. The pore size is
chosen dependent on the molecular range to be investigated.
Since all analytes are eluted between the exclusion volume and the total hold-up volume
of the column only a limited number of components can be separated. Therefore, it is
important to choose a column with matrix particles having pore sizes suitable for the
molecules in the sample to be separated. A given SEC column has a given molecular weight
range for fractionation. If the fractionation range for proteins is given as 30 000 to 200 000, it
means that proteins larger than 200 000 are excluded from the pores and thus they all elute
with the exclusion volume. All substances less than 30 000 fully penetrate all the pores, get
the maximum retention and elute together in the hold-up volume (the total permeation
volume). The column can thus separate substances with molecular weights between 30 000
and 200 000. Thus, large molecules that do not enter the pores cannot be separated from each
other and analytes that are so small that they all penetrates all pores cannot be separated.
Packing materials can be soft or hard. Soft materials are made of polysaccharides such as
dextrans, polyacrylamide or polystyrene. Soft gels are used with gravity flow (flow driven
only by gravity), where the mobile phase flows through the column as a function of gravity
only. For HPLC rigid packing materials that resist the higher pressure are used. Rigid
packaging materials are made of silica or a highly crosslinked organic polymer like
160
Introduction to Pharmaceutical Chemical Analysis
polystyrene–divinylbenzene copolymer. Silica is used with aqueous mobile phases, while
polystyrene–divinylbenzene often is used with organic solvents as mobile phases.
When molecular weight is to be determined by SEC it is important to realise that the size
of large molecules like proteins or polysaccharides depend on the environment they are
dissolved in. A coiled molecule may change it size due to its concentration, to the ion
strength of the solvent, to solvation and so on. Calibration standards with known molecular
weight have to be of a similar structure as the analytes to be determined. Otherwise there is a
risk of a major bias in the result. The calibration curve is a logarithmic function and small
variations in elution volume lead to major errors in the molecular weight determined. Thus
the accuracy of the molecular weight determination by SEC is not impressive and the
technique is therefore more suited for the characterization of polymers than for actual mass
determination. The selectivity in SEC can be varied by changing the pore size distribution of
the column packing material.
11.6.2
Summary of SEC
Size exclusion chromatography is used for the separation of substances by their size in solution.
The stationary phase consists of porous particles with a controlled pore size. Large molecules
are excluded from the pores and the volume of mobile phase transporting them out of the
column is called the exclusion volume. Small molecules penetrate the pores and are retarded on
the column. The volume of the mobile phase transporting them out of the column is called the
total permeation volume. SEC can be used for separation of substances with molecular sizes
corresponding to elution volumes in the window between the exclusion volume and the
permeation volume. Peak capacity is limited compared to other separation techniques.
11.7 Ion Exchange Chromatography
Ion exchange chromatography is a technique that allows the separation of ions based on their
charge. It can be used for almost any kind of charged molecules, including large proteins,
nucleotides and amino acids. The ionic analytes are retained by ionic interaction between
the analytes and ionic sites of the opposite charge placed on the stationary phase. Ionic
groups like –SO32, –COO, –NH3þ or –NR3þ are placed on the matrix of the particles. The
charges are neutralized by counter ions which can exchange with analytes.
The positive cations have an affinity to negatively charged ion exchangers and these ion
exchangers are therefore called cation exchangers, and positively charged ion exchangers
are likewise pronounced anion exchangers. In daily work a strong cation exchanger is
abbreviated SCX, and a strong anion exchanger is abbreviated SAX. Similarly weak ion
exchangers are pronounced WCX or WAX.
The ion exchanger can be classified into strong and weak ion exchangers, where the
strong exchangers have either sulfonic acid groups, –SO32, or quaternary ammonium
groups, –NR3þ, attached to the stationary phase and the weak exchangers have carboxylic
acid groups, –COO, or an aliphatic amino group, –NH3þ, attached. The aliphatic amino
group can be a primary, secondary or tertiary amine. The strong ion exchangers are charged
throughout the usable pH range while the weak exchangers are only charged within a minor
range of the pH interval depending on the pKa value of the functional group.
Chromatographic Separation Principles
161
The ionic adsorption sites on the column packing material are very specific and therefore
the column packing materials have a given capacity. The capacity is between one and four
milli equivalents (meq) per gram. A capacity of 4 meq/g of cation exchanger means that 1 g
maximum can bind 4 mM of a protonated tertiary amine. It is important to realize that it is
not possible to use the full capacity of a column when samples are injected. Only amounts of
analytes of 5–10% of the capacity should be loaded onto such a system per injection in order
to avoid overloading of the stationary phase.
Retention of analytes on an ion exchanger requires that the analyte as well as the ion
exchanger are charged. In general the affinity of analytes depends on their charge and size.
More charges and a smaller hydrated size increase the affinity for the ion exchanger and
thus retention.
The retention can be changed by modifying the mobile phase either by the addition of an
organic solvent or more effectively by changing the concentration or nature of the buffer.
The buffer ions in the mobile phase compete with the analyte ions for the ionic sites on the
stationary phase. When the buffer concentration is increased or buffer ions with a stronger
affinity to the ionic sites are used the retention of the analytes decreases. In cation exchange
the eluting strength of the buffer cation increases in the order Liþ, Hþ, Naþ, NH4þ, Kþ, Agþ,
Mg2þ, Zn2þ. In anion exchange the eluting strength of buffer anions increases in the order
OH, CH3COO, HCOO, Cl, Br, H2PO4, oxalate, citrate. This is of course only a
general rule and may change with the type of ion exchanger. It is common to perform
gradient elution in ion exchange chromatography by increasing the ionic strength of the
buffer during chromatography.
12
Thin-Layer Chromatography
Thin-layer chromatography (TLC) is a liquid chromatography separation technique
where the stationary phase is dispersed as a thin layer placed on the surface of a plate
typically made of glass or alumina foil. Typical applications are the identification of
active ingredients and contaminants in pharmaceutical raw materials and preparations
and the analysis of herbals used as drugs or food supplements. The samples, prepared in
a volatile solvent, are placed as spots or bands on the plate parallel to one of its edges.
The sample solvents are then evaporated and the plate inserted in a tank with a mobile
phase at the bottom of the tank. The development tank is closed and the mobile phase
migrates up the plate by capillary forces. The analytes in a spot migrate up the plate
depending on their distribution between the mobile phase and stationary phase. When
the mobile phase has moved a certain distance towards the top of the plate, the TLC
plate is removed from the tank and the mobile phase is evaporated from the plate in a
fume hood. The position, color and size of the spots in the sample seen on the plate are
now compared with the position, color and size of spots in reference substances. More
details are given in this chapter.
12.1 Introduction
Thin-layer chromatography (TLC) is the simplest of the chromatographic separation
techniques because it does not require additional apparatus for the mobile phase to
move through the stationary phase. TLC in its most simple form is very inexpensive
and easy to perform. The technique is fast because many samples can be placed on the TLC
plate simultaneously. TLC has a wide scope and stationary phases for all liquid chromatography separation principles are commercially available. TLC has for many years been the
technique of choice for impurity testing of drug substances and finished products in the
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
164
Introduction to Pharmaceutical Chemical Analysis
pharmacopoeias. The use of the TLC technique for the characterization and quality
assessment of herbals is also an important area of application.
Normal phase chromatography using the polar silica material as stationary phase and
organic solvents as mobile phases is often the first choice, but reversed phase chromatography using aqueous mobile phases is becoming increasingly popular. The substances are
made visible on the plate with general or specific detection methods. General visualization
techniques can be used to view all the analytes, while more specific methods of detection
may be used for the visualization of substances with specific properties. In addition to TLC
as being a simple and inexpensive technique, it also displays all the components in the
sample, including substances with a strong interaction to the stationary phase, and with no or
only a short distance of migration. When using HPLC and GC, it is easy to overlook
substances strongly retained because, if they elute at all, they may elute very late as broad
peaks with very long retention times.
The limitation of TLC is primarily related to the sensitivity of the technique. Detection
limits are significantly higher than for other separation methods such as HPLC, GC and CE.
Small, volatile substances cannot be analyzed because they may evaporate with the solvent
when spotting the sample on to the TLC plate or when removing the eluent after
development.
12.2 Apparatus
In its simplest form, not much apparatus is needed to perform TLC analysis. Figure 12.1
shows the apparatus and procedure.
A small volume (e.g., 10 ml) of the sample solution is deposited on the TLC plate using
micropipettes along a starting line approximately 2 cm from the edge. The distance between
each spot or band must be at least 1 cm as some band broadening takes place during
chromatography. The sample solvent is evaporated, and the TLC plate is then placed in a
tank containing the mobile phase. Following the chromatographic separation the TLC plate
is removed from the tank, the mobile phase is evaporated and the substances on the TLC
plate are identified. Colored analytes are directly visible, and substances that absorb UV
light or fluorescence can be detected when the TLC plate is irradiated with light from a UV
lamp. The TLC plate can also be sprayed with reagents that react with substances to form
colored compounds.
Retention of the substances on the TLC plate is described by the retention factor, Rf. The
Rf value is defined as:
Rf ¼
a
b
ð12:1Þ
where a is the distance from the starting line to the center of the spot, and b is the distance
between the starting line and the front line of the mobile phase as shown in Figure 12.2.
Rf values will be in the range between 0 and 1 and analytes with the strongest interaction
with the stationary phase are retarded most and have the lowest Rf values. In tables of
retention factor Rf values are often multiplied by 100 so that the values given are between 0
and 100.
Thin-Layer Chromatography
165
Micro
pipette
Starting
line
•
•
•
Mobile phase
Solvent
front
•
•
TLC
spray
•
Figure 12.1 Apparatus and procedure for TLC. Apply the sample to the TLC plate, perform the
chromatography in the development tank, then identify the spots on the TLC plate
Solvent front
b
a
•
Starting line
Figure 12.2 Parameters for determination of Rf values
166
Introduction to Pharmaceutical Chemical Analysis
12.3 TLC Plates
The thin layer of stationary phase is coated onto plates or thin sheets typically of glass,
plastic or alumina. TLC plates can be made in the laboratory, but their quality is often less
good than commercially available TLC plates, and in pharmaceutical quality control TLC
plates of high and reproducible quality are needed. Two types of TLC plates are available,
differing in the particle size of the stationary phase:
.
.
Normal TLC plates;
High performance TLC plates (HPTLC plates).
Normal TLC plates has a layer of stationary phase 0.25 mm thick and with particle sizes of
5–12 mm. The plates are typically 20 20 or 20 10 cm. In high performance thin-layer
chromatography (HPTLC) plates stationary phases having smaller particle sizes of 2–10 mm
are used and the layer thickness is only 0.1–0.2 mm. The size of the plates is typically
10 10 cm in size or half the size of normal TLC plates.
The use of smaller particles with a narrower particle size distribution as well as the
reduced plate size in HPTLC results in the same (or even better) separation compared to
normal TLC plates but in a much shorter time and using less mobile phase. The reduction in
particle size, as predicted by vanDeemter (see Chapter 13), increases the separation
efficiency. As the spots or bands are much narrower on the HPTLC plates lower limits
of detection can also be obtained. More narrow spots or bands have to be used on HPTLC
plates in order not to compromise the separation efficiency, and applied sample volumes of
less than 1 ml are often used.
If a larger volume of sample is to be applied to the TLC plates this can be achieved with a
so-called concentration zone. The application area for the sample solution consists then of a
material that does not provide strong interaction with the analytes and that has the same
porosity as a layer of stationary phase. When the TLC plate is placed in the developing
chamber with the mobile phase, the liquid moves the substances in the concentration zone to
the edge of the stationary phase, where they are concentrated as a thin band. Kiselguhr
(diatomaceous earth) is often used as the material in the concentrating zone. Kiselguhr is the
same material used as a support material in packed columns for GC.
In Ph.Eur. the methods prescribed for the identification of herbals allow a free choice
between the use of normal TLC and HPTLC plates. At present (2010) a major effort is made
to transform tests using TLC for the analysis of related substances in active pharmaceutical
ingredients (API) into HPLC methods. The HPLC methods provide a more reliable
estimation of the quantities of possible impurities present in the APIs.
12.4 Stationary Phases
The most common stationary phase for TLC is silica gel or silica gel-based materials but
also alumina (aluminium oxide) is sometimes used, and cellulose may be a choice for more
polar analytes.
The silica gel used is similar to the one used in HPLC (see Chapter 13) and may therefore
be used in the normal phase mode (unmodified silica gel) as well as in the reversed phase
mode (e.g., silanized silica gel).
Thin-Layer Chromatography
167
The stationary phase may also be impregnated with a buffer or another substance to
facilitate a certain separation. It is important to be aware that the silica gel easily picks up
water and other compounds from the atmosphere and therefore it can be necessary to clean
the plate, for example, by pre-elution with methanol and subsequent heating to above 100 C
to remove the adsorbed water.
In order to make the layer of stationary phase more robust a binder can be added during
manufacture. Previously gypsum (calcium sulfate) was used and silica gel treated in this
way was designated G. However, organic polymers like polyvinyl alcohol provide a much
more robust layer that can withstand most handling in the laboratory. These TLC plates are
therefore preferred by most laboratories.
Silica used for stationary phases can be chemically modified similar to what has been
described for HPLC and in this way hydrophobic stationary phases are generated. Silica
with octyl or octadecyl hydrocarbon chains bonded to the surface is common and is used
with relatively polar mobile phases which are typically water mixed with methanol or
acetonitrile. This type of chromatography is called reversed phase mode.
12.5 Mobile Phases
Mobile phases for TLC are volatile liquids and are removed after chromatography.
Therefore the spectroscopic characteristics like UV transparency have no influence on
the subsequent detection like it has in HPLC. Since the plates are used only once, there are
no pH limitations as there is with HPLC columns. Most applications are based on normal
phase chromatography in which silica is the stationary phase and an organic solvent or a
mixture of organic solvents is used as the mobile phase. Retention is determined by the
nature of the analytes and of the solvent strength of the mobile phase. Analytes with very
polar groups or having the greatest number of polar groups have the highest affinity to the
silica and are therefore retarded the most. The solvent strength of the mobile phase is
adjusted to provide optimum retention on the plate so that Rf values are in the range 0.2 to
0.8. Low solvent strength provides strong retention (small Rf values) and high solvent
strength results in low retention (large Rf values). In normal phase chromatography the
solvent strength is given by the e value of the solvent or solvent mixture, as shown in
Chapter 11. Solvent strength increases with the value of e . Pentane has an e value of 0 and
is the weakest solvent in normal phase mode, while methanol has an e value of 0.95 and is a
strong solvent. It is very common to use mixtures of solvents to obtain a suitable solvent
strength and selectivity of separation. For example, the USP recommends the mixture of
chloroform/methanol/water (180 : 15 : 1 v/v) as a general mobile phase used when no other
mobile phase is specified for the test.
There may be strong ionic interactions between amino groups and the acid silanol groups
on the stationary phase. The result is that analytes with amino groups are more strongly
retarded, and in addition, the spots often show tailing on the plate. This can be counteracted
by adding a basic component like ammonia or a small volatile amine like diethylamine to the
mobile phase. The basic component then competes with the analytes for the acid silanol
groups. The interaction can also be suppressed by adding a volatile acid such as formic acid
or acetic acid to the mobile phase. In this way the ionization of the silanol groups can be
suppressed and the ionic interaction decreases. Carboxylic acids are also strongly retarded
168
Introduction to Pharmaceutical Chemical Analysis
by silica. To make them migrate further and more uniformly up the plate acetic acid or
formic acid can be added to the mobile phase.
In the reversed phase mode chemically bonded silica is used for the stationary phase. In
this mode the apolar/hydrophobic analytes are retarded more strongly than the polar
analytes. Mobile phases are aqueous and therefore only organic solvent miscible with
water can be used. Methanol/ethanol and acetonitrile are obvious choices but also acetone
can be used. Acetone has a lower boiling point than methanol and acetonitrile, and it
evaporates faster. Increasing the content of organic modifier in the mobile phase reduces
the retention and provides increasing Rf values. To adjust the retention of ionized analytes
volatile acids or bases such as formic acid, acetic acid and ammonia or salts like
ammonium acetate can be added to the mobile phase. Buffers such as phosphate buffers
are normally not employed.
12.6 Chromatographic Development
Normally TLC plates are placed vertically in a development tank as shown in Figure 12.1.
This is called vertical development. The mobile phase is added to the development tank to
about 1 cm above the bottom and the tank is covered with a lid. The inner walls of the tank
can be covered with filter paper that sucks up the mobile phase. This enables a rapid
equilibrium between the mobile phase and the vapor phase in the tank. Equilibration
takes about 1 h.
To initiate the development place the TLC plate in the development tank as shown in
Figure 12.1. Keep the tank open for as short a time as possible to avoid disturbance of the
saturation in the tank. If the equilibrium is destroyed it can have an influence on the
reproducibility of the Rf values which normally increase if the development is performed in
a unsaturated tank. The TLC plates are removed from the tank when the front of the mobile
phase has migrated between two-thirds and three-quarters of the total length of the plate. In a
fume hood the mobile phase is now evaporated from the plate using a hair dryer and the
detection of the separated sample can now take place.
Analyzing very complex samples using a simple development as described above does
not always give sufficient separation. A two-dimensional approach can then be used
(Figure 12.3). After the first development the dried plate is rotated 90 so that the partly
separated sample is just above the surface of the new mobile phase.
The technique is called two-dimensional development. The mobile phase used in the
second dimension is optimized to separate the substances that were not separated by the
first mobile phase. Of course only one sample (spot) can in this way be analyzed per
TLC plate.
In most tests, when using normal TLC plates, vertical ascending development is applied.
HPTLC plates can be developed in vertical ascending mode as well as in a horizontal mode.
The horizontal mode requires a special development chamber (Figure 12.4).
The mobile phase migrates faster by horizontal development, and the analysis time is
thus shorter. The TLC plate is positioned so that the layer of stationary phase forms one of
the walls of the chamber which is then saturated with vapors from the mobile phase. The
volume of the chamber is considerably smaller than the volume of a normal development
tank. Therefore the equilibration time is much shorter, and the reproducibility of
separations increases.
Thin-Layer Chromatography
Run in first dimension
169
Run in second dimension
Figure 12.3 Two-dimensional TLC
Top cover
Reservoir with
mobile phase
Mobile phase controlling
chamber saturation
TLC plate oriented with
stationary phase downwards
Figure 12.4 Development chamber for horizontal development of HPTLC plates
12.7 Detection
After chromatographic development and after the mobile phase has been removed by
evaporation, the substances are detected. Colored analytes are immediately visible. To
facilitate visualization of the noncolored spots or bands on the TLC plate after chromatography a phosphor can be added to the stationary phase during the manufacturing process.
Silica gels treated in this way are often designated F254 as the plates fluoresce when viewed
under UV light of 254 nm (using a mercury lamp). Analytes absorbing UV light of this
wavelength appear as dark spots on a greenish fluorescent background (fluorescence
quenching). For analytes that fluoresce themselves TLC plates without the phosphor
incorporated is preferred. If the analytes do not absorb UV light or if an extra selectivity
in detection is preferred the developed plates can be treated with reagents for derivatization
and subsequently heated if necessary, forming colored spots or bands. In Ph.Eur. a number of
reagents are described and a huge range of such reagents are available in the TLC literature.
12.8 Applications of TLC
In Ph.Eur. TLC is used for the identification of drug substances and herbals and in tests for
known and unknown contaminants. Solutions of samples, reference substances and a
170
Introduction to Pharmaceutical Chemical Analysis
Figure 12.5 TLC spots with the same Rf value, representing different sample concentrations
suitability test mixture are applied to the TLC plate and after development the dried plate is
examined. The following parameters are compared:
.
Does the suitability test comply with requirements?
Compare the spot retention of the reference substance and the sample.
. Compare the spot size and color of the reference substance and the sample.
.
The suitability test is included to secure that the chromatographic system has the sufficient
separation capability. In the identity test the spots (Rf value, color and size) for analyte and
reference substances have to be identical.
In the limit tests for impurities the spot sizes on the TLC plate and the spot intensities are
compared. Analytes with different Rf values in principle do not have the same size due to the
increased band spreading with increasing Rf values. It can therefore be difficult to obtain a
reliable result when spots with very different Rf values are compared. Spots with the same Rf
value, however, have been exposed to the same band spreading and should therefore have the
same size if the initial spots were of the same size.
However, when viewing spots of the same substance but representing different amounts
the spot size appears to correlate with the amount. But in principle the spots are of the same
size, it is only a question of what the human eye is capable of detecting. The eye is not able to
see the small nuances of color that the low drug concentration provides. Since drug
concentrations are highest in the center of the spot and decrease towards the edge of the
stain, the stains with low drug concentrations are less than those with higher drug
concentrations. This is illustrated in Figure 12.5.
12.9 Quantitative Analysis and Instrumentation
Semi-quantitative limit tests for impurities are used in the pharmacopoeias and in these tests
it is necessary to compare the spot size of the detected impurity with an authentic reference
substance in a suitable concentration in order to be able to deliver a reliable test answer. In
this way it is typically possible to detect impurities down to 0.5–1.0% of the active
substance. For documentation it is common to obtain a photograph of the TLC plate.
0.2
0.4
0.6
Rf
0.8
SPD
TYR
CAD
171
HIS
NH3
HIS
TYR
PUT
NH3
AU
CAD
AU
SPD
PUT
Thin-Layer Chromatography
5
10
15
20
min.
Figure 12.6 Comparison of the analysis of biogenic amine by TLC (left) and HPLC (right)
A number of instruments have been developed in order to improve the reliability of TLC
methods. Apparatus for the application of sample solutions is available and can be
recommended for the application of spots or band to HPTLC plates. For quantitative
analysis densitometry is available.
The densitometer scans the TLC trace in either UV absorbance mode, UV quenching
mode or fluorescence mode. The detector signal obtained as a function of Rf value is similar
to a chromatogram (Figure 12.6) and quantification can be performed based on peak area
measurements which is more reliable than the human eye.
Recently, instrumentation for the hyphenation of TLC with mass spectrometry has
become available. This improves the possibility to identify unknown substances detected on
the TLC plate.
12.10 Summary
TLC is a technique which is very inexpensive and easy to perform with no or very little
requirement for instrumentation. TLC has for many years been one of the most important
methods in the pharmacopoeias for identity testing and for testing for known and unknown
contaminants in drugs, drug substances and herbals. In most tests normal TLC plates is
used with vertical development. HPTLC plates are used when normal TLC plates do not
provide enough separation or if a reduction in analysis time is important. The stationary
phase is typical of silica and the chromatography is performed using volatile organic
solvents or mixtures thereof as eluents. This is the principle of normal phase (straight
phase) adsorption chromatography where the most polar substances have the strongest
retention and the lowest Rf values. TLC plates with silanized silica are used for reversed
phase separation. In this case the mobile phase is typically a mixture of water or aqueous
172
Introduction to Pharmaceutical Chemical Analysis
buffer with methanol or acetonitrile. Polar substances are retarded the least and have the
highest Rf values. Noncolored substances are visualized by derivatization to colored
products or by irradiation of phosphor-containing TLC plates with UV light to give dark
spots on a fluorescent background.
For limit tests and quantification the Rf value, size and color of the spots are compared
with the spots of reference substances. This can be done manually or by the use of
instrumental methods.
13
High Performance Liquid
Chromatography
High performance liquid chromatography(HPLC, or for short LC) is the most commonly
used chromatographic technique to determine drugs in pharmaceutical preparations and in
biological material. In LC the mobile phase is a liquid, forced through a column packed with
a material that retards the analytes introduced into the system. The analytes are injected into
the flow of mobile phase just in front of the separation column. The outlet of the column is
connected to a detector where the eluted substances are detected. The separation principle
can be any of the principles described in chapter 11. This chapter reviews how the apparatus
is constructed and how it works.
13.1 Introduction
The names of LC are many. When HPLC was introduced in the late 1960s the name high
pressure liquid chromatography was used. The pressure is however not a desire but more an
unavoidable drawback, and the name was change to high performance liquid chromatography, focussing on the very good separations obtained.
It is now becoming common practice just to use the shortened LC. Systems capable of
running at very high pressures have also been introduced under the name of ultrahigh
performance liquid chromatography (UHPLC or just UPLC).
The main components of an LC system are shown in Figure 13.1.
The separation columns are typically made of steel tubes 5–25 cm long and packed with
the stationary phase. It is in the column that substances are separated and thus the column is
the heart of the chromatograph. The other parts of the chromatograph are optimized
individually and carefully put together to optimize the separation efficiency of the total
system. The three main parts of the LC system: the solvent delivery, the separation column
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
174
Introduction to Pharmaceutical Chemical Analysis
HPLC
pump
Pre column
Sample
injection
Analytical
column
Filter
Column
oven
Mobile
phase
reservoir
Detector
Data system
Waste
Figure 13.1 Main structure of an LC system
and the detector are all vital and indispensable units. The separation occurs when the mobile
phase is pumped at a constant flow rate through the column bringing the separated analytes
to the detector. In a standard analytical HPLC system the typical flow rate of the mobile
phase through the column is 0.5–2.0 ml/min but can be in the range from 0.01 to 10 ml/min.
The small particle size of the column packing materials result in a back pressure of
30–300 bar (3–30 MPa) and in certain cases (UHPLC) up to 1200 bar (120 MPa), when the
mobile phase is pumped through. The pumps used must be able to pump the mobile phase at
a constant flow rate against a high pressure. Any particle in the samples injected collects at
the top of the column and gradually blocks the column, with an increase in back pressure as
the result. This results in a decrease in flow rate that compromises the analytical result.
Therefore HPLC pumps are equipped with a regulation mechanism that keeps the flow rate
constant, and a gradual blockage thus results in an increase in back pressure while the flow is
kept constant.
The substances to be separated must be dissolved in a liquid that is miscible with and not
stronger eluting than the mobile phase. Typical injection volumes are 5–100 ml. The
injection systems are optimized to inject the solution under high pressure directly into
the flowing mobile phase just before the column inlet. The detectors provide an electronic
response to the analytes. The response is processed by a computer system that prints the
results as chromatograms. The whole analysis process can be automated and controlled by
the computer system. When an autosampler is used as the injector the LC system can work
24 h daily. Analytical chemical data, including quantitative calculations can be reported
continuously by the computer system.
The broad applicability and the high degree of automation of HPLC are among the
reasons why the technique has gained such a dominant position in analysis of food,
High Performance Liquid Chromatography
175
pharmaceuticals and biofluids. HPLC is important in drug research and development
because:
.
HPLC with UV detection provides accurate, precise and robust methods for the
quantitative determination of drugs in pharmaceutical preparations.
. HPLC is very suitable for monitoring the stability of drug substances and drugs in drug
preparations and for quantifying decomposition.
. HPLC is very well suited for the determination of drugs and their metabolites in biological
material.
13.2 The Chromatographic Separation Process
The dynamics of a chromatographic separation process apply for all the different chromatographic separation techniques: TLC, HPLC, GC and SFC. In order to obtain a basic
understanding of which parameters influence the quality of the separation performed and
consequently to be able to optimize a chromatographic system a short introduction to this
field shall be given.
The molecules in a sample are exposed to a number of physical actions on their transport
through the chromatographic system. The molecules are introduced in the mobile phase and
interact with the stationary phase during the chromatographic process. One of the important
parameters for the result of the separation is the flow rate of the mobile phase. In the 1950s
van Deemter and coworkers were already studying the effect of the mobile phase on the
efficiency of the GC separation and they could express the efficiency in this formula – the
van Deemter equation:
H ¼ Aþ
B
þ Cu
u
ð13:1Þ
This formula expresses the correlation between the efficiency given by H, the height
equivalent to a theoretical plate, and the band broadening phenomena A, B and C as a
function of the flow rate, u.
The contribution to the A term comes from the fact that identical molecules travel
different distances on their way through the chromatographic system due to the non-uniform
packing of the particles of the stationary phase (Figure 13.2). The contribution of the A term
to H for a given chromatographic system is more or less independent of the flow rate.
The B term takes account of the diffusion of the molecule (Brownian movements) in the
mobile phase. Drug molecules dissolved in a liquid always diffuse away from areas in
the liquid with a high drug concentration towards areas of lower concentration until the
concentration of the substance is the same throughout the fluid volume. The analyte
molecules diffuse in all directions and the radial diffusion does not affect the efficiency. But
the diffusion in the direction of the length of the column, the longitudinal diffusion, results in
band broadening especially if the flow rate is low. The B term becomes insignificant at
higher flow rates. The B term is of higher significance in GC than in HPLC due to higher
diffusion at higher temperatures and to less hindrance of diffusion from other mobile phase
molecules.
176
Introduction to Pharmaceutical Chemical Analysis
Initial
analyte
band
width
°°°°°°°° °° °° °° ° ° °°
°°
°°°
°°°°°
START
°°°°
°
°°°°
Final
analyte
band
width
EDDY DIFFUSION
Figure 13.2 Peak broadening due to eddy diffusion
The C term covers the mass transfer between the two phases. This involves the direct back
and forth transfer of molecules between the mobile and the stationary phases (Figure 13.3).
A peak broadening results as the drug molecules at a given time period retarded in the
stationary phase are stationary while the drug molecules in the mobile phase are not retarded
and thus pushed ahead by the mobile phase.
The van Deemter equation can also be visualized as shown in Figure 13.4
The van Deemter equation has been reworked by many researchers and a more detailed
equation used for HPLC could be:
H ¼ Ce dp þ
Cm dp2 u Cd Dm Csm dp2 u Cs df2 u
þ
þ
þ
Dm
Dm
Ds
u
ð13:2Þ
where Dm and Ds are the diffusion in the mobile phase and in the stationary phase,
respectively, dp is the particle diameter of the particles constituting the chromatographic
bed and df is the layer thickness of the stationary phase. Ce, Cm, Cd, Csm and Cs are
constants. In this equation special consideration has been given to diffusion and mass
transfer in the mobile as well as in the stationary phases (Figure 13.5). From this it is
obvious that the particle diameter, dp, is of major importance and the plate height
decreases when the particle diameter decreases. A smaller plate height gives room for
more plates in a given column length and the efficiency thus increases.
Flow
Equilibrium
concentration
Mobile phase
Actual
concentration
Stationary phase
Figure 13.3 Mass transfer between mobile and stationary phase
High Performance Liquid Chromatography
177
H
Minimum
of H
=
optimum
efficiency
C
B/u
A
u
Optimum of u
Figure 13.4 The van Deemter plot showing the optimum efficiency (at minimum H) and the
optimum flow rate
Mobile mass transfer takes place as the flow rate through channels is different in the
middle compared to close to the side walls of the channels where the flow rate approaches
zero. Both phenomena give rise to a broadening of the analyte into a larger volume.
Mass transfer in the stationary phase also involves the transfer to stagnant pools of
stationary phase in closed pores of the porous particles.
13.3 The Column
The column is the heart of the separation process and therefore it is important to keep it in
good condition at all times. It does not help to have a D 40 000 instrument if the column is
unable to separate analytes.
The column should have a high number of theoretical plates, N, and as known from the
van Deemter equation, this depends on the particle size of the column packing material.
°
°° °°
°
°° °° °
°
Band
width
°°
°°
Mobile phase
mass transfer
°°
Stagnant mobile phase
mass transfer
°°
Band
width
°° Band
°°
Stationary phase
mass transfer
Figure 13.5 Mass transfer in mobile and in stationary phase
width
178
Introduction to Pharmaceutical Chemical Analysis
HETP (μm)
10 μm
2.0
5 μm
3.5 μm
HPLC
1.5
1.0
UHPLC
1.5 μm
2
3
4
Linear velocity
5
6
7 mm/s
0.5
1
Flow rate (ml/min)
ID = 2.1 mm
0.15
0.3
0.45
0.6
0.75
0.9
1.05
ID = 4.6 mm
0.7
1.4
2.1
2.8
3.5
4.2
4.9
Figure 13.6 Height equivivalent to a theoretical plate as a function of the flow rate of the mobile
phase for 1.5, 3.5, 5, and 10 mm particles
Decreasing particle size provides a larger number of theoretical plates. This is illustrated in
Figure 13.6 where the height equivalent to a theoretical plate (HETP) is plotted versus
particle sizes of 1.5, 3.5, 5 and 10 mm.
The optimum efficiency is at the smaller particle sizes shifted towards higher linear flow
rates. This indicates that higher flow rates can be used without losing efficiency. However,
the small particle size also provides a very high back pressure and thus shorter column
lengths have to be used.
The practical implication of this is that short columns with particles of about 2 mm can
give efficient separations in a very short time (Figure 13.7, Table 13.1). Depending on the
back pressure it can be necessary to use UHPLC equipment.
The typical analytical HPLC column has been a 15–25 cm long steel tube packed with
5 mm particles. Inner diameter of the tube has been 4.6 mm. With the introduction of mass
spectrometry as a routine LC detection technique and the improvement of column packings
for HPLC, column dimensions of 10–15 cm with 2 mm internal diameter have become the
common standard. For fast analysis smaller particles between 1.5 and 2.0 mm is used in
columns of 3–5 cm in length, and this is designated as UHPLC.
Using shorter columns with smaller internal diameters results in major savings in the
consumption of mobile phase. In Table 13.2 the reduction in mobile phase consumption is
calculated for columns of equal length with different internal diameters for a constant flow
rate of mobile phase through the columns.
High Performance Liquid Chromatography
179
0.20
AU
Column: 150 mm
Particle: 5 μm
Flow rate: 0.2 ml
0.10
0.00
0.00
2.00
4.00
6.00
10.0
8.00
Minutes
0.20
Column: 100 mm
Particle: 3.5 μm
Flow rate: 0.3 ml
AU
0.10
HPLC
0.00
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
4.0
Minutes
0.20
Column: 75 mm
Particle: 2.5 μm
Flow rate: 0.5 ml
AU
0.10
0.00
0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80
2.0
Minutes
0.20
Column: 50 mm
Particle: 1.7 μm
Flow rate: 0.6 ml
AU
UHPLC
0.10
0.00
0.00
0.20
0.40
0.60
0.80
1.00
1.0
Minutes
Figure 13.7 Effect of the reduction in particle size and column length and the analysis time and
chromatographic separation
Table 13.1 The correlation between column length, efficiency N (resolving power) and
particle size
Column length
(mm)
Resolving power, N
Particle size
5.0mm
3.5mm
1.8mm
150
12 500
21 000
—
100
75
8500
6000
14 000
10 500
—
—
50
30
15
4200
—
—
7000
4200
2100
12 000
6500
2500
180
Introduction to Pharmaceutical Chemical Analysis
Table 13.2 Saving in mobile phase as a result of the reduction in internal column diameter
keeping the linear flow rate (1.33 mm/s) in the column constant
Internal diameter (mm)
Flow rate (ml/min)
Reduction in mobile phase (%)
1.0
0.69
0.43
0.19
0.05
—
31
57
81
95
4.6
4.0
3.0
2.0
1.0
A further reduction of consumption in mobile phase is proportional to the simultaneous
reduction in column length. A reduction in column length also results in a similar reduction
in analysis time. The column packing material is held in place in the column by a metal filter
at each end. The metal filter is either a porous frit or a net so that the mobile phase can pass
through. The pore size of the filter has to be smaller than the diameter of the packing material
in order to prevent particles from leaking from the column.
A narrow particle size distribution of the column packing material is important for a good
and uniform packing of the column. This, combined with low dead volume end fittings of the
column and short tubing with low internal diameter from the injection device to the column
and from the column to the detector, secures high quality of the separations. If non-optimal
connections of tubings and fittings are made when installing the column in the LC system,
the separation can be less efficient due to extra column band broadening even if the column
itself is in perfect condition. Remember, when measuring the efficiency of the “column” it is
the total system that is being tested.
13.4 Pumps
The pumps deliver the mobile phase at a constant flow rate through the column. The pumps
can be constructed in different ways, but a piston pump is the most common. Figure 13.8
shows a sketch of a piston pump.
The piston pump consists of a small steel cylinder with a volume of approximately 100 ml.
A piston is moved back and forth in the cylinder by means of a motor. There are two ball
To HPLC
column
.........
.
Motor
Pulse
dampener
Check
valves
Piston
.........
Mobile phase
inlet
Figure 13.8 The main parts in a piston pump
High Performance Liquid Chromatography
181
valves attached to the cylinder so that the mobile phase can only flow in one direction, into
the cylinder from the reservoir or out to the column. When the piston is moved back the
lower ball valve opens while the upper ball valve closes, dragging the mobile phase into the
cylinder. When the piston again is moved forward into the cylinder the bottom ball valve
closes, while the top valve opens. The mobile phase is then forced out of the cylinder through
the injector into the column. Since the mobile phase is forced into the column only when the
plunger is pushed into the cylinder, the fluid flow pulsates. This pulsation introduces extra
noise in the detector signal and should be eliminated if possible. A pulse dampener is
therefore included in the system to ensure a smooth flow of the mobile phase. Other pump
systems ensure a smooth flow by linking together two piston heads into a double piston
pump, where one piston head delivers mobile phase to the column while the second is filled
up with mobile phase.
When the pumping system delivers a mobile phase with a constant composition to the
column it is called isocratic elution. However, it is also possible to have a system where
liquid is pumped from two or more reservoirs and then to mix them during chromatography.
This can be done using a single pump equipped with a low pressure mixing valve connected
to up to four different reservoirs containing liquids. The mixing valve opens for only one
pipeline at a time and in this way the solvent mixture can be controlled. It is also possible to
use more than one pump where each pump delivers a controlled amount of each liquid. The
mixing is then performed at the high pressure side, but the two principles are of similar
performance.
The change of the composition of the mobile phase during chromatography is called
gradient elution and this technique can be compared to temperature programming in GC.
Gradient elution is used to separate samples containing compounds with large differences
in retention. When big differences in retention of analytes are current, isocratic elution
results in unnecessary long analysis time. The least retarded substances elute early as
partially overlapping peaks, while the most retarded substances elute as broad peaks with a
long retention time. This is illustrated in Figure 13.9.
It is only the analytes in the middle of the chromatogram that elute satisfactory. Using
gradient elution the composition of mobile phase is changed during chromatography,
starting with a weaker eluting composition of the mobile phase. In this way the least retarded
substances obtain sufficient retention and separation. The strength of the mobile phase is
then increased during chromatography and the late elution peaks are now eluted faster and
with a better sensitivity. This is illustrated in Figure 13.10.
Figure 13.9 Isocratic elution of a sample containing analytes with large differences in retention
182
Introduction to Pharmaceutical Chemical Analysis
Gradient
profile
Figure 13.10 Gradient elution of a sample containing analytes with large differences in
retention
13.5 Detectors
The LC detector gives a response for the analyte that is converted into an electrical signal.
The response should be proportional to the concentration of either substance in the mobile
phase or with the mass of the substance in the mobile phase, so quantitative analysis can be
carried out based on the measurement of peak areas or peak heights. The detectors can be
divided into two types: (i) general detectors that measure any change in the mobile phase and
(ii) specific detectors that respond only to substances with specific properties. The main
detectors used in pharmaceutical analysis are given in Table 13.3.
For the quality control of pharmaceutical preparations and drug substances the UV
detector is the standard detector. Fluorescence detectors and electrochemical detectors have
for selected analytes a much lower limit of detection than the UV detector. The mass
spectrometer provides additional information on the molecular structure. Mass spectrometers are treated elsewhere in this book (Chapter 16). A refractive index detector is used in
quality control in the case of substances that do not have an adequate UV response. It is
considerably less sensitive and it is not applicable to methods using gradient elution.
13.5.1
UV Detector
The UV detector is based on the analytes’ absorption of UV light. Analytes capable of
absorbing UV radiation can be detected. This requires the analyte to contain a chromophore,
being at least one double bond in the molecule. The wavelength range is from 190 to about
Table 13.3 Some commercial available LC detectors and their typical performance
Detector
Ultraviolet (UV)
Fluorescence
Electrochemical (ECD)
Mass spectrometry (MS)
Refractive index (RI)
Evaporative light scattering (ELSD)
Charged aerosol (CAD)
Lower limit of detection (ng)
Gradient elution?
0.1–1.0
0.001–0.01
0.01–1.0
0.001–0.01
100–1000
0.1–1.0
0.1–1.0
Yes
Yes
No
Yes
No
Yes
Yes
High Performance Liquid Chromatography
183
400 nm, and at higher wavelengths the visible range up to about 800 nm can be used. Colored
analytes are more selectively detected in the visible region although they also absorb light in
the UV region. It is first and foremost the high operational stability and ease that makes the
UV detector the preferred detector for quality control. It is also very good when using
gradient elution. The lower limit of detection is adequate for most analysis, but it is not
sufficiently sensitive to be used for the analysis of low concentrations of drugs in biological
material. Figure 13.11 shows a sketch of a UV detector.
The eluent from the column is directed to a flow cell through which also UV light of a
defined wavelength is directed. The radiation which passes through the flow cell is detected
by a light sensor.
It is common to use a deuterium lamp as radiation source. A continuum of light is emitted
in the whole UV range and the detector can therefore be used continuously variable in the
range 200–400 nm. For optimal detection sensitivity of the substances they should be
measured at their maximum UVabsorbance. A monochromator ensures that UV radiation of
the correct wavelength is directed through the flow cell.
The measurement of absorbed radiation is according to Beer’s law proportional to the
concentration of the substance in the mobile phase, to the path length of radiation through
the flow cell and to the absorptivity of the substance. The absorbance is proportional to the
radiation path through the flow cell.
Figure 13.12 shows a sketch of a flow cell. The eluent from the column flows through a
Z-shaped channel in the cell. The UV radiation passes the flow cell through two quartz
windows that do not absorb UV radiation. The path length of the flow cell is in the range from
6 to 10 mm and the volume for a standard cell is in the order of 10 ml. When very efficient
separation is obtained it can be necessary to use a flow cell with a smaller internal volume in
order to avoid mixing of the separated peaks in the cell.
In the diagram of the UV detector shown in Figure 13.11, the monochromatic radiation
passes through the flow cell and is directed towards the detection unit. In another type of
UV detector called a diode array detector (DAD) the polychromatic radiation is passed
Light
sensor
Flow
cell
Slit
Grating
Lamp
Slit
Figure 13.11 Schematic diagram of an UV detector
184
Introduction to Pharmaceutical Chemical Analysis
Mobile phase
to waste
Slit
Lamp
Quartz
window
Light
sensor
Mobile phase
from column
Figure 13.12 Sketch of flow cell in a UV detector
through the flow cell. After the flow cell the transmitted light is split in an optical lattice
into the individual wavelengths, and the intensity of each of these is measured by a number
of (array of) photodiodes. There may be up to several hundred diodes in series to measure
the intensity of the array of UV radiation. Diode array detectors offer several possibilities.
A full UV spectrum of the peak can be recorded “on the fly”, which can be used to identify
the substance. It is also possible to choose a few selected wavelengths, so that each
substance in a sample can be detected at the optimal wavelength.
13.5.2
Fluorescence Detector
All substances that are fluorescent also absorb radiation, but it is far from all substances that
absorb radiation that fluoresce. The fluorescence detector is more selective than the UV
detector. It also has a lower limit of detection than the UV detector, but this is only true for a
selected number of analytes. For such compounds it can be used to detect low concentrations
of substances in biological samples.
Figure 13.13 shows a schematic diagram of a fluorescence detector. The lamp is usually a
xenon lamp emitting an intense continuum of light in the whole UV region and in parts of the
visible region.
The wavelengths of the excitation and emission (fluorescence) radiation are controlled by
monochromators. The molecules are excited using an intense light beam of a selected
wavelength (the exitation wavelength). Some of the absorbed energy is released by heat
dissipation and the rest of the energy is emitted as light having a longer wavelength (the
emission wavelength) than the excitation wavelength (Figure 13.14). The excitation
wavelength can be chosen from the UV spectrum of the compound. Normally it is preferred
to excite the compound at the wavelength where it has its maximum UV absorption but for
High Performance Liquid Chromatography
185
Slit
Flow
cell
Lamp
Slit
Grating
Slit
Light
sensor
Figure 13.13 Schematic diagram of a fluorescence detector
selectivity reasons another wavelength can be chosen. Also the emission wavelength chosen
for measuring can be varied as the emission spectrum also covers a range of wavelengths.
For dilute solutions, the following equation expresses how the fluorescence is dependent
on a number of parameters:
F ¼ F Io a b c
ð13:3Þ
Fluorescence intensity F is proportional to the quantum yield F, the intensity of the
excitation radiation Io, the absorbance of the compound a, the path length of the detector b
and the concentration of the compound in the mobile phase c. The quantum yield is a
compound-dependent constant in a given environment.
Fluorescence is emitted in all directions but is normally measured at a 90 angle to the
excitation radiation. This reduces background noise and in principle the fluorescence is
Intensity
Excitation spectrum
= absorbance
Emission spectrum
= fluorescence
HO
N
H3CO
N
300
450
600 nm
Figure 13.14 Excitation and emission spectra of quinine
186
Introduction to Pharmaceutical Chemical Analysis
measured on a dark background. This is also why the fluorescence detector has a lower limit
of detection than the UV detector.
Since fluorescence intensity is proportional to the intensity of the radiation that excites the
molecules of the substance, an increase in radiation intensity provides a proportional
increase in fluorescence intensity.
13.5.3
Electrochemical Detector
Electrochemical detection is a selective detection principle for compounds with electrochemically active groups that can be either reduced or oxidized. Oxidation is in practice
much easier to perform than reduction because the reduction mode requires that oxygen is
totally removed from the mobile phase. The oxidation is performed at a given voltage
typically between þ 0.3 and þ 1.0 V. The higher the voltage the more the substances are
oxidized. The detector measures the current as a function of oxidation (Figure 13.15). The
detector is especially well suited for the detection of phenols, amines and thiols.
The detector is not as robust as the UVand fluorescence detectors, and the electrodes can
be contaminated by impurities in the mobile phase and in the samples. Pulsed amperometric
detection expands the applicability of electrochemical detection to cover alcohols and
carbohydrates. In this mode carbohydrates can be detected with high sensitivity, but it has to
be done at high pH (pH 12–13) and this of course puts some extra demands on the
HPLC system.
13.5.4
Refractive Index, Evaporative Light Scattering and Corona
Discharge Detectors
A sensitive, universal detector is a chromatographer’s dream. Refractive index, evaporative
light scattering and corona discharge detectors are detectors that to a certain limit meet
the dream. The refractive index (RI) detector measures the change in the refractive index of
the mobile phase leaving the column. An analyte with a refractive index different from the
mobile phase is detected. However, the detector has low sensitivity and is easily influenced
by changes in temperature. In the evaporative light scattering detector (ELSD) the column
effluent is nebulized using nitrogen at an elevated temperature. Nonvolatile compounds
form an aerosol that scatters the light from a diode laser. Volatile substances cannot be
measured by this technique.
O
O
R
HO
R
HO
2 e-
Figure 13.15 Oxidation of catecholamines release electrons that are detected by the electrochemical detector
High Performance Liquid Chromatography
187
The charged aerosol detector (CAD) uses a similar principle for detection. The effluent is
evaporated and an aerosol of nonvolatile compounds is formed. Nitrogen molecules passing
a corona receive positive charges which are transferred to the aerosol particles. The
positively charged aerosol particles are then measured by an electrometer. Both ELSD
and CAD are universal detectors with the exception of volatile substances.
13.5.5
Combination of Detectors
It is possible to combine detectors in series and the combination of a UV detector, a
fluorescence detector and an electrochemical detector in series is a powerful tool. The
detectors have to be coupled in the order given above. The fluorescence detector cell does
not withstand higher backpressures and the electrochemical detector has to be placed last in
the series as no backpressure is allowed. The major advantage of this set-up is that the
relative detector signals achieved for a given compound in the three detectors should be
constant for a reference standard and for the similar peak in a complex sample. If this is not
the case there probably is some interference in the sample that biases the analytical result.
When also a MS detector is to be included in such a series it is necessary to split the eluent
after the UV detector with one part going to the MS and the other part to the fluorescence and
electrochemical detectors.
13.6 Injectors
The purpose of the injector is to bring a certain volume of sample solution into the flow of
the mobile phase just before it enters the top of the column. The high back pressure of the
column necessitates a special devise for this purpose. An injector that is leak proof in the
whole pressure range of the system is needed. For manual injections a simple loop injector as
shown in Figure 13.16 can be used.
Sample
injection
From
pump
From
pump
To waste
Loop
To HPLC
column
To waste
To HPLC
column
Loop
Figure 13.16 A loop injector in the load position (right) and in the inject position (left)
188
Introduction to Pharmaceutical Chemical Analysis
The loop injector is a six-port valve. In the load position the mobile phase from the
pump passes through the valve directly to the column. In this position it is possible to
inject the sample into the loop using a syringe. When the valve then is switched to the
inject position, the mobile phase from the pump passes through the loop and in this way
brings the sample to the column. The loop is a piece of tubing with a total volume which
should not exceed the capacity of the column used. In standard analytical work it is
common to use a loop of 20 ml together with column of 4.6 mm internal diameter. There
are two ways to load the sample into the loop, either by partial filling where, for example,
a volume of 5 or 10 ml is injected into the loop, or by filling the whole loop with sample.
In the latter case it is important to flush the loop with a sample volume of five times
the loop volume in order to remove the previous solvent (mobile phase) completely from
the loop. After an injection the injector is switched back to the load position. Channels
can be flushed with a cleaning solution to remove residual sample solution. It is
preferable to use a full loop injection technique in order to achieve the highest
repeatability of the injected volume.
The injection process can be automated using an autosampler. Samples from sample vials
placed in the autosampler are injected into the HPLC according to the pre-programmed
volumes and time intervals. The process can be controlled by the HPLC software so that
analysis can take place without supervision, running for 24 h or more. Autosamplers can
also be equipped with a refrigerator for the cooling of sensitive samples so that they do not
decompose before analysis.
13.7 Mobile Phases
The following requirements for liquids to be used for mobile phases have to be considered:
.
The solvents should not give any response in the detector used.
The solvents must have a satisfactory degree of purity. A number of solvents are available
in so-called HPLC quality and sometimes specified for selected purposes (e.g., for
gradient elution or for MS detection).
. The solvents should have low viscosity to provide as low a back pressure as possible in the
HPLC system.
. The solvents should have low toxicity, preferable be inflammable and nonreactive, and
they must be suitable for disposal after use.
.
Mobile phases for reversed phase chromatography often contain aqueous buffer solutions.
The buffer salts are dissolved in water of HPLC quality. Water of HPLC quality is produced
in the laboratory using a water treatment plant that removes common contaminants from tap
water, or it can be obtained commercially. The aqueous solution is mixed with an organic
modifier to the prescribed solvent strength. If the buffer salts give rise to particles in the
solution it has to be filtered through a filter with pore size of 0.45 mm. Before use, the mobile
phase has to be degassed to remove dissolved air. Degassing can be performed by ultrasound
treatment, by purging with helium or by vacuum treatment. Vacuum treatment is most
common and is now included in most instruments as a continuous online degassing system.
Degassing is necessary in order to prevent bubble formation in the detector cell causing
noise in the detector signal.
High Performance Liquid Chromatography
189
Table 13.4 UV cutoff for common solvents (1 cm path length)
Solvent
Acetone
Acetonitrile
Dichloromethane
Ethanol
Heptane
Methanol
Tetrahydrofuran
Water
UV cutoff (nm)
330
190
220
210
195
205
210
190
The mobile phase is contained in a bottle. The pump sucks liquid from the bottle through
a polyethylene tubing equipped with a filter that prevents any particles from entering
the pump.
The UV detector is by far the most widely used HPLC detector. Analytes can be detected
when they have a UV response above 195 nm. This requires the mobile phase to be
transparent at the given wavelength. Table 13.4 gives an overview of the UV cutoff of
common solvents.
The UV response increases as the wavelength decreases, and the UV cutoff value is the
shortest wavelength that can be used with the solvent.
13.8 Solvents for Sample Preparation
A prerequisite for the samples to be analyzed by HPLC is that they are soluble in the
mobile phase and the mobile phase is therefore the first choice as a solvent for the sample.
In general, the substances should not be dissolved in a solvent with a higher solvent
strength than the mobile phase. The use of 100% methanol should be avoided, because the
sample solution then has a too high solvent strength. If substances dissolved in neat
methanol are injected, the methanol prevents the substances from being retarded on the
stationary phase when they enter the column. The substances remain in the methanol plug
until it has been diluted with the mobile phase and equilibrium has been re-established in
the system. The result is that the substances are eluted from the column earlier and as
broad peaks. If methanol must be used as a solvent, it must be injected only in small
volumes in order to prevent peak broadening. If larger sample volumes are to be injected,
it is preferable to have a lower solvent strength in the sample compared to the mobile
phase. The substances can then under the correct conditions be concentrated on top of the
column during the injection and extra bands spreading because of dilution in the mobile
phase can be avoided.
13.9 Reporting the Results
The signals from the detector are collected either by a PC with suitable software or by an
integrator. The results are printed out as chromatograms. In addition the software reports the
190
Introduction to Pharmaceutical Chemical Analysis
retention time, the peak height and the peak area of each peak in the chromatogram. The
software can also do calculations and the results of quantitative analysis can be calculated
and printed as part of the analysis report. Many HPLC instruments are supplied with a PC
holding software that controls the entire system, including the injection of the samples from
the autosampler. These HPLC systems are being considered analysis robots which can
analyze samples and report results 24 h daily.
13.10 Summary
A liquid chromatograph consists of a pump that delivers the mobile phase from a reservoir
and onto a column packed with a column packing material that typically consists of 3–5 mm
silica particles with a hydrophilic or hydrophobic surface. Sample solutions are injected
using a pressure- and leak-resistant injector into the mobile phase just before the column. In
isocratic elution the mobile phase has the same composition throughout analysis while in
gradient elution the solvent strength of the mobile phase is increased during analysis. The
substances eluted from the column are detected by one or more detectors. The UV detector is
the standard detector in pharmaceutical quality control. In addition, detectors with a lower
limit of detection (e.g., the fluorescence detector, the electrochemical detector, the mass
spectrometer) are used for determination of drug substances and their metabolites in
biological material.
14
Gas Chromatography
Gas chromatography is a separation technique used for the analysis of volatile substances
or substances that can be made volatile. Analytes are separated as a function of different
distribution ratios between the mobile and the stationary phase. The mobile phase is a gas
under pressure flowing through the heated column containing the stationary phase. In gas
chromatography the mobile phase is called the carrier gas. The stationary phase is placed
on the inner surface of the column or coated onto particles packed into the column.
The column is a long tube made of fused silica or metal. The sample is injected into a heated
injector at the beginning of the column where the constituents evaporate. The carrier gas
carries the analytes through the column where the substances are distributed between the
stationary phase and the mobile phase. Analytes with different distribution ratios between
the phases are carried through the column at different speeds towards the detector. The
separated substances are detected by a detector placed at the end of the column. Gas
chromatography is suitable for the separation of substances which are volatile and
thermally stable or can be converted into volatile and thermally stable derivatives.
14.1 Introduction
In GC the mobile phase is an indifferent gas that does not react with the analytes. Analytes
must be volatile in order to be transported by the carrier gas to the detector. Consequently,
only volatile substances can be separated by GC. The volatility of the substances is
temperature-dependent, the higher the temperature the higher the vapor pressure of the
substances. Substances that are readily volatile and evaporate easily can be analyzed at a
lower temperature than less volatile substances. The temperature can be too high, because
the substances may decompose at high temperatures. The applicability of GC as to which
substances that can be analyzed is therefore dependent on the volatility and thermal stability
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
192
Introduction to Pharmaceutical Chemical Analysis
Figure 14.1 Photograph of a gas chromatograph
of the substances in question. Large polar molecules such as peptides, proteins, nucleotides,
polysaccharides and other biomolecules cannot be separated by GC.
14.2 Apparatus
Figure 14.1 shows a photo of a gas chromatograph and Figure 14.2 shows an outline of the
principle.
The carrier gas is delivered from a high pressure steel cylinder and is delivered through a
reduction valve and into the GC instrument. The carrier gas is heated and passes through
temperature controlled flow controllers on to the injector. The heated carrier gas flows
through the injector through the column, to the detector and out to waste.
Flow
control
Injection
port
Detector
Carrier gas
Data system
Capillary column
Column oven
Figure 14.2 Outline of a gas chromatograph: gas cylinder, reduction valve, flow controllers,
injector, column, detector, data collection on PC
Gas Chromatography
193
The reduction valve brings the pressure from the cylinder down from 200 bar to about
5 bar. The flow controllers are used to make sure that a constant inlet pressure is maintained
and thus a controlled flow through the column. This is important if repeatable retention
times are to be obtained. The substances are usually dissolved in a volatile solvent and
injected with a syringe into the GC. Common injection volumes are 0.5–2.0 ml. The
temperature of the injector is kept high enough to evaporate the substances immediately,
and the carrier gas brings the evaporated sample on to the column. A detector is placed at the
end of the column and detector signals are collected by computer software. The computer is
also used to control the whole GC system.
The GC column is placed in an column oven with a fan that ensures good circulation and
temperature control during the analysis. The detector is heated to a temperature that is higher
than the temperature of column oven to avoid condensing of substances in the detector when
leaving the column. The exhaust from the detector should somehow be led to the central
ventilation system. In the column the substances are distributed between the carrier gas and
the stationary phase. The retention decreases with increasing column temperature because
the vapor pressure of the substances increases and the distribution is driven towards the gas
phase.
The stationary phase is either a nonvolatile liquid or a solid. A solid stationary phase is an
organic or an inorganic polymer and separation is obtained due to either different adsorption
to the surface of the polymer or by sieving through well defined pores of the polymer. This
technique is named gas–solid chromatography (GSC) and is primarily used to separate
substances that are gases at normal room temperature or are small, very volatile organics.
The use of a nonvolatile organic liquid as stationary phase is named gas–liquid chromatography (GLC). This is the dominating separation principle in GC analysis. Separation is
based on the distribution of substances between the gas and the stationary liquid. The main
parameters that control a GC separation are:
.
Temperature of injector, column and detector;
Type and amount of stationary phase;
. Column dimension;
. Type and speed of carrier gas.
.
14.3 Temperature
The temperature of the column is the main parameter that controls retention in GC analysis.
The substances are distributed between the gas and liquid phases, but they only move
forward through the column when they are in the gas phase. An increase in temperature
increases the volatility of the substances and in this way changes the distribution favoring
the gas phase. Therefore, reduced retention is observed with increasing column temperature.
As a rule of thumb, the retention time is halved when the column temperature is increased
about 30 C. The column temperature is generally lower than the boiling point of the
substances. Column temperatures above 300 C increase the risk of decomposition of
analytes and are rarely used.
The temperature of the injector, the column and the detector are carefully controlled in
each GC method and the temperature is optimized for the analysis of a given type of sample.
Introduction to Pharmaceutical Chemical Analysis
194
C9 C10 C11
C13
C12
Isothermal at 150ºC
C14
C7
C8
C15
0
10
20
30
45
55
90
100 min.
C9 C10 C11 C12 C13
Temperature
programming
from 50ºC to 250ºC
C14
C8
C15
C16
C7
C17
0
10
20
30
C18
C19
40 min.
Figure 14.3 Isothermal versus temperature programming in GC analysis of paraffins
The temperature at the injection site is normally higher than the column temperature and so
high that the substances evaporate very fast. If the injection temperature is too low it takes
longer to bring analytes into the gas phase, resulting in broader peaks and poorer separation.
Too high an injection temperature can lead to decomposition of compounds during
injection. The detector temperature is generally higher (25–50 C higher) than the column
temperature. This can prevent substances with low volatility from condensing in and
contaminating the detector.
Samples containing few constituents not too different in volatility can be separated using
a constant column temperature. This is called isothermal GC analysis. If samples containing
analytes with a wide span in volatility are analyzed by isothermal GC, analytes with high
volatility have short and overlapping retention times while analytes with low volatility elute
as broad peaks with long retention times (Figure 14.3). The separation of such complex
samples is improved using temperature programming of the column temperature, starting at
a relatively low temperature and increasing to a sufficiently high temperature during the
chromatographic process. A temperature increase of 2–10 C/min is common. Most GC
analyses are performed by temperature programming. The outcome of temperature
programming is comparable to the outcome of gradient elution in HPLC.
Gas Chromatography
195
14.4 Carrier Gas
The carrier gas is an inert transport medium for the substances. Gasses like nitrogen, helium
and hydrogen can be used. Short analysis times are desirable, and this is achieved by using a
high gas velocity through the column. However, it turns out that the peaks become broader
both if the gas speed is too low as well as if it is too high. Highest efficiency is achieved by an
optimum gas velocity. The optimum gas velocity is different for nitrogen, helium and
hydrogen. This is shown in Figure 14.4 where the height equivalent to a theoretical plate
(HETP) is plotted against the mean gas velocity through the column. Figure 14.4 shows the
average linear gas velocity through the column (cm/s), versus HETP (mm). This is called a
van Deemter plot (see Chapter 13).
Nitrogen provides an optimum gas velocity at a lower flow rate, while hydrogen is shown
to have an optimum gas velocity at a higher flow rate and consequently results in a shorter
analysis time with similar efficiency. Used at the optimum gas velocities, the analysis time is
more than halved by using hydrogen as a carrier gas compared to the use of nitrogen. When
used at flow rates higher than the optimum it is important to note the steepness of the van
Deemter plots. Increasing the flow rate above the optimum results in a significant loss of
separation efficiency when using nitrogen compared to hydrogen (and helium). The
steepness expresses the readiness in mass transfer between the mobile and stationary
phases. The nitrogen molecule is relatively large and analyte molecules are more likely to
bump into nitrogen molecules than into the small hydrogen molecules. This implies that
nitrogen hampers mass transfer, resulting in broader peaks. Helium is safer to use than
hydrogen and is therefore preferred as a carrier gas. Analysis time is reduced by 40%
compared to nitrogen.
N2
HETP (mm)
0.8
He
H2
0.4
20
40
60
80
Carrier gas velocity (cm/s)
Figure 14.4 van Deemter plots for different carrier gasses
196
Introduction to Pharmaceutical Chemical Analysis
CH3
CH3
CH3
H3C Si O Si O Si CH 3
CH3
CH3 nCH3
CH3
CH3
H3C Si O Si O
CH3
CH3
CH3
CH 3
CH 3
Si CH3 H 3C Si O Si O Si CH3
CH3 (CH 2)3 CH 3
CH3
CN
p
x
Figure 14.5 The basic skeletons of polydimethylsiloxane, polyphenylmethylsiloxane, and
polycyanopropylmethylsiloxane
The consumption of carrier gas using capillary columns is in the range 1–10 ml/min. For
packed columns the consumption is significantly higher, 40–60 ml/min. Nitrogen is
preferred as carrier gas for packed columns because it is safe to use, inexpensive and
readily available. The mobile phases in GC do not influence the selectivity of the separation,
which is very different from the mobile phases in HPLC. However, a high purity of the
carrier gas with a low content of oxygen and organic impurities is important in order to avoid
deterioration of the stationary phase and high background in the detector response.
14.5 Stationary Phases
Stationary phases in GC are temperature-stable liquids having very low vapor pressure.
Many different (hundreds) have been suggested through the years, but if GC columns are to
be used at temperatures up to or even above 300 C, the choice of appropriate substances is
limited. Only a number of polymers possess the characteristics to be liquid at or close to
room temperature, to be stable and to have a low vapor pressure at high temperature.
The preferred stationary phases are therefore either polysiloxanes or polyethylene glycols.
The polydimethylsiloxane polymer shown in Figure 14.5 is the most hydrophobic of the
polysiloxanes. The polarity can be varied by substituting the methyl groups for phenyl or
cyanopropyl groups. Thus a number of stationary polysiloxane phases exist with increasing
number of phenyl groups, providing increasing polarity of the stationary phases. When
cyanopropyl groups are used the polarity is increased further.
Polysiloxanes are very temperature-stable and can be used over a wide temperature range.
The introduction of the polar cyanopropyl groups reduces the temperature stability from
being above 300 C to about 275 C.
Polyethylene glycol represents another family of stationary phases. The general structure
is shown in Figure 14.6.
Polyethylene glycols are also named Macrogol or Carbowax followed by a number
indicating the molecular weight. The temperature stability increases and the vapor pressure
decreases with increasing molecular weight. These polymers are available with molecular
weights in the range from 300 to 107, but the most popular polymer for GC is Carbowax 20M
having a molecular weight of 20 000. It can be used in the temperature range from 60 to
250 C. It is of course important to be aware of the upper temperature limit in order to avoid
HOCH 2 CH2 O CH2 CH2 OH
m
Figure 14.6 The basic skeleton of polyethylene glycol
Gas Chromatography
197
column bleeding (loss of stationary phase) and degradation of the stationary phase, but it is
also important to use a column temperature above the lower limit. Below the lower
temperature limit the stationary phase becomes too viscous and the diffusion of the
analyte is reduced, leading to broader peaks. The hydroxyl group terminating the polyethylene glycol chain is polar and the stationary phase is characterized as being of
intermediate polarity. It also provides a different selectivity towards analytes compared
to polysiloxsane phases.
14.6 Selectivity in GC
As mentioned previously the retention of substances is governed by the temperature of
the column. Increasing the column temperature increases the volatility of the substances
and the retention decreases. An increase in column temperature of 30 C as a rule of
thumb bisects the retention time. The solubility of the analyte into the stationary phase is
another important parameter influencing the retention. The solubility depends on the
interaction that takes place between the substances and the liquid phase. Strong
intermolecular interactions lead to increased solubility. The types and strength of
intermolecular forces can be seen in Table 3.2 Intermolecular interactions are needed
for a substance to be dissolved into another. Water cannot be dissolved in heptane as the
water molecules prefer to interact with other water molecules through hydrogen
bonding. The heptane molecules cannot participate in hydrogen bonding and stick
together through van der Waals forces. Heptane and water have similar boiling points, 98
and 100 C respectively. If they were to be separated in GC on an apolar stationary phase
like polydimethylsilicone the water molecules would pass through the column unretained with the carrier gas, while the heptane would have a retention due to its solubility
in the apolar stationary phase. If the stationary phase was changed to a more polar one
like a carbowax phase, water molecules can now interact through hydrogen bonding with
the stationary phase and are retained, while the heptane molecules only show very little
or no retention on this phase.
A very nonpolar stationary phase provides little interaction with most substances, while
the interaction becomes stronger as the polarity of the phases increases. The interactions can
be dipole–dipole interactions or hydrogen bonding interactions. Dipole–dipole interactions
are most important when using silicone polymers as stationary phases, while hydrogen
bonding interactions are more dominant when using polyethylene glycol stationary phases.
When using the apolar stationary phase polydimethylsiloxane, neither dipole–dipole
interactions nor hydrogen bonding interactions are in force, and the retention of solutes is
determined by their volatility. This leads to the following rule:
On apolar stationary phases analytes are separated according to their boiling points,
and analytes with low boiling points are eluted before substances with higher boiling
points.
The introduction of phenyl groups in the stationary phase increases the retention of
substances due to an increase in polarity. The phenyl groups give rise to intermolecular
198
Introduction to Pharmaceutical Chemical Analysis
Table 14.1 Order of elution for a mixture of ethanol, 2-butanone, benzene, and cyclohexane
chromatographed on three different stationary phases. The boiling points are given: 1 elutes
before 2, etc
Stationary phase
Polydimethylsiloxane
Polymethylphenylsiloxane
Polyethylene glycol
Ethanol,
78.5 C
2-butanone,
79.6 C
Benzene,
80.1 C
Cyclohexane,
81.4 C
1
1
4
2
2
2
3
4
3
4
3
1
interactions due to temporary dipoles. Retention increases further with an increase in the
percentage of phenyl groups and/or cyanopropyl groups in the stationary phase due to an
increase in dipole–dipole interactions. Polyethylene glycol also provides hydrogen bonding
interactions with analytes, especially compounds having an alcohol functional group, and
alcohols have a relatively stronger retention than other analytes. This leads to a general rule
for theselection of a stationary phase expressed as “like dissolves like”. In practice this
means that the choice of stationary phase has to be tuned to be similar in polarity to the
analytes to be separated. Retention depends on both the substances’ boiling point and
solubility in stationary phase.
The consequence is that different stationary phases provide different selectivity, and the
order of elution of analytes can thus be changed by changing the stationary phase. An
example of this is shown in Table 14.1.
On polydimethylsiloxsane the substances are eluted according to their boiling points.
Ethanol has the lowest boiling point and elutes first. The difference in retention time is small
because the difference in boiling points is small. Benzene has a relatively higher solubility in
polymethylphenylsiloxane due to p–p interactions and consequently has a relatively
stronger retention on this phase, while ethanol is the most soluble in polyethylene glycol
and thus has the strongest retention in this phase.
14.7 Columns
In GC two different types of columns are used:
.
.
Capillary columns;
Packed columns.
The two types of columns differ that much in dimensions that they are not interchangeable
on the same GC instrument. Originally GC instruments used packed columns but today most
GC separations are performed using capillary GC instruments.
14.7.1
Capillary Columns
Capillary columns can be made of a number of materials like metal or fused silica. The fused
silica capillary is preferred due to its robustness and inactivity. Glass columns are fragile and
Gas Chromatography
199
metal columns may contribute to analyte decomposition. The outer surface of the fused
silica capillary is coated with a polyimide layer to improve the mechanical stability.
The columns are very flexible, and they are easy to handle in daily practice. The columns are
similar to the capillaries used in capillary electrophoresis (CE), but the dimensions are
different. The inner diameter is in the range 20–500 mm, typically 250 or 320 mm. The
stationary phase is a thin film on the inside of the column. It can either be physically coated
on the inner surface or be chemically bonded to the surface. The more stationary phase
placed in the column, the greater is the retention. Film thickness increases when the amount
of stationary phase increases. Film thickness is in the range from 0.05 to 10.0 mm, and
0.25 mm is a typical film thickness. Thin film thickness is used in the analysis of heavy
volatile substances as it reduces retention and provides acceptable analysis times, without
using too high column temperatures. Thicker films are used to increase the retention of more
volatile substances.
The major difference of the capillary column compared to the packed column is that the
capillary is an open tube, and there is only a little pressure drop across the column when the
carrier gas flows through. Therefore it is possible to use fairly long capillary columns
without the pressure becoming too high. In pharmaceutical applications, the typical column
length is in the range 10–30 m. As a rule of thumb the number of theoretical plates (N) on a
capillary column is about 3000 per meter. A column length of 30 m will therefore provide
about 90 000 theoretical plates. This makes GC on capillary columns one of the most
powerful separation techniques. It is not unusual that one can detect 50–100 peaks or even
more in a chromatogram.
14.7.2
Packed Columns
The packed columns are – as the name indicates – packed with particles that are covered
with the stationary phase. The particles make up the solid support for the stationary phase
and are therefore called the carrier support. The supporting material must be inert and should
have a surface that provides a smooth, uniform film of stationary phase. Particle size is
usually 100–200 mm. Smaller particles result in more dense packing of the columns and the
back pressure increases. This is avoided by using larger particles. Column lengths of up to
2–3 m are common.
The most common support materials are made of diatomaceous earth, which is the
skeleton of single-celled siliceous diatoms. The solid support is purified and pretreated
to make it suitable for GC. The amount of stationary phase on the support material is
given as a percentage, and the normal range is 1–10%. Thick layers of stationary phase
are used to provide readily volatile substances with sufficient retention as well as for
preparative purposes. To avoid high column temperatures when analyzing heavy
volatile substances columns with a thin layer of stationary phase is used. It is common
that the columns are made of glass, but columns made of metal are also used and the
internal diameter is in the range 2–5 mm. The packing material causes back pressure in
the column. Therefore, packed columns are much shorter in length than capillary
columns. The number of theoretical plates in a packed column is approximately 2000
per meter.
The particles give rise to different path lengths of the analyte molecules through the
column. This results in peak broadening (cf. van Deemter; see Chapter 13).
200
Introduction to Pharmaceutical Chemical Analysis
14.8 Injection Systems
In GC, it is common to inject the substances dissolved in a volatile solvent. The Solvent and
the substances evaporate in the injector and the gas mixture is brought to the column by the
carrier gas. The chemicals are injected with a syringe, and regular injection volumes are in
the range 0.5 to 2.0 ml. It is important that the sample evaporates instantly and that the
volume brought to the column does not overload the column. Otherwise peak broadening
will occur.
Injection systems for capillary columns and packed columns are different. This is because
the capillary is much narrower in internal diameter and thus contains less stationary phase
than a packed column. Thus the risk of overloading the stationary phase increases with
decreasing amounts of stationary phase. Injection systems for capillary columns are
designed to eliminate overloading during injection.
14.8.1
Injection Systems for Capillary Columns
A 1-ml sample solution in the injector forms, after evaporation, about 0.5–1.0 ml of gas. The
injector must therefore have a sufficiently large volume to allow the sample solution to
expand to this gas volume. Moreover, the injection system must prevent the formed gas from
entering directly into the capillary column. The internal volume of the capillary is so small
that 1 ml of gas fills large parts of the column, preventing separation.
In order to prevent overloading of the column three different injection systems are
available:
.
Split injection;
Splitless injection;
. Cool on-column injection.
.
Split injection and splitless injection use the same injector design. Figure 14.7 shows a
principle sketch of a split/splitless injector.
The sample solution is injected with a syringe into the injector. The syringe needle
penetrates a silicon membrane (septum), and the tip of the syringe needle is in this way
inserted into a glass tube in the injector. The sample solution is injected and evaporates in the
glass tube. The inner volume of the glass tube is large enough to allow expansion of the
sample into a gas.
The principle of the split injection is that the evaporated sample is split so that only a small
portion of the sample is brought from the injector into the column, while the rest is vented
out of the injector. The heated carrier gas enters the injector which has three outlets. The GC
capillary column is attached to one outlet. Another outlet is just under the septum to prevent
any impurity from the septum reaching the column. The third outlet is called the splitter
outlet and this outlet is used to vent the sample volume that is not brought to the column.
The gas flow rates through the three outlet are controlled, and it is the relationship between
the gas velocities through the column and the splitter output that determines how much
sample volume is brought into the capillary. Assume a flow of carrier gas of 2 ml/min
through the column and 50 ml/min through the splitter output. In this case the sample
volume is split in the ratio 1 : 25, and only 4% of the sample enters the column, while 96% is
Gas Chromatography
201
Septum
Carrier gas
inlet
1-100 ml/min
Septum
purge
0.1-2 ml/min
Liner
Split
purge
1-100 ml/min
Analytical
GC column
1 ml/min
Figure 14.7 Split/splitless injector for capillary GC
vented through the splitter output. Split ratios are normally in the range from 10 : 1 to 100 : 1.
Since only a small portion of the sample volume is used for the separation and detection,
split injection is best suited for the analysis of sample solutions containing relatively high
concentrations of analytes. Split injection is common in analysis related to the quality
control of raw materials and drug preparations.
The use of split injection for the analysis of complex samples containing analytes with
large differences in volatility may involve the risk of analyzing a nonrepresentative sample
fraction. This can be the case if decomposition of some analytes occurs at the high
temperature of the injector or if some analytes of low volatility condense in the injector
before they reach the column.
In splitless injection the total sample volume is injected into the column. The principle is
that the sample evaporates in the injector and is delivered by the carrier gas to the top of the
column where the analytes are concentrated. In this mode the splitter output is closed during
injection. If a 1-ml sample expands to 1 ml of gas and the carrier gas velocity through the
column is 2 ml/min, it takes 30 s to transfer the sample to the column. To prevent spreading
of the analytes throughout the capillary during the 30 s of transfer time of the sample, the
analytes must be concentrated at the column entrance. There are two ways to perform this
concentration:
.
.
By solvent effect;
By cold trap.
A solvent effect is obtained if the solvent used for dissolving the analytes condenses in the
very first part of the column, forming a thin film. The analytes can then be trapped in the
202
Introduction to Pharmaceutical Chemical Analysis
condensed solvent. The temperature of the column has to be about 20–50 C below the
boiling point of the solvent to obtain the effect. After the analytes have been trapped in
the solvent the column temperature is raised by temperature programming and the
solvents and substances then evaporate. With this technique low limits of detection can
be obtained because the full injection volume is used for detection. In the cold trap
technique a powerful cooling system is used to cool the top of the column. This
technique is best suited to trap very volatile substances and has found little application in
pharmaceutical analysis.
A third injection technique is direct injection on a cold column. As the name says, this is
an injection of the sample solution directly into the capillary without a preceding
evaporation. The column temperature is sufficiently low that the solvents do not evaporate.
The sample is injected through a valve into the column using a syringe with a very fine
fused silica needle thin enough to enter into the column. Since the sample solution is
brought directly into the column, it should not contain nonvolatile contaminants.
Nonvolatile matter may block the column. The technique is rarely used in pharmaceutical
analysis.
14.8.2
Injection Systems for Packed Columns
Packed columns have a higher capacity due to their higher amount of stationary phase and
the whole sample volume can be injected into the column. The sample solution is injected
through the resealable silicone–rubber septum directly into the upper part of the column
placed in the injector (Figure 14.8) where it evaporates and then is transferred to the column
packing by the carrier gas.
Septum
Carrier gas
inlet
Empty part of
GC column
Liner
Packed
Analytical
GC column
Figure 14.8 Injector for packed columns
Gas Chromatography
203
Exhaust
Cylindrical
collector
electrode
Anode (+)
Cathode (-)
Insulations
H2
Air
From
GC column
Figure 14.9 Schematic drawing of a flame ionization detector
14.9 Detectors
There are a large number of detectors available for GC, but only the standard detectors
used in pharmaceutical analysis are described in detailed here.
14.9.1
Flame Ionization Detector
In the flame ionization detector (FID) the analytes enters a flame and are ionized, and ions
are formed in the detector house. The rise in the electrical current corresponding to the
number of ions formed is measured between two electrodes. Figure 14.9 shows a sketch of
the detector.
The carrier gas leaving the column is mixed with hydrogen at the entrance to the detector,
while the air is led into the detector house. The carrier gas/hydrogen mixture containing the
analytes enters the detector house through a jet. The flame is burning on the top of the jet and
when the organic analytes burn in the flame ions and electrons are formed. The amount of
charged particles is measured by applying a voltage of 300 V between the jet and the
collecting electrode just above the flame. The current measured is proportional to the
amount of organic matter that burns.
FID is the most commonly used GC detector. It gives a linear response for organic
compounds over a wide concentration range. The lower limit of detection is about 10–9 g.
The FID detector is standard in most GC analysis in Ph. Eur. and USP.
14.9.2
Nitrogen–Phosphorus Detector
The nitrogen–phosphorus detector (NPD) is an ionization detector that provides a selective
response for organic substances containing nitrogen or phosphorus. The detector is also
204
Introduction to Pharmaceutical Chemical Analysis
Exhaust
Anode
Cylindrical
collector
electrode
Heater
Rubidium or
Cesium bead
Cathode
Flame
Insulations
H2
Air
From
GC column
Figure 14.10 The nitrogen–phospherous selective detector
called an alkali flame ionization detector (AFID) or thermionic detector (TID). The detector
is built up similar to the flame ionization detector, but with a crystal of alkali metal salt
placed just above the flame (Figure 14.10). The alkali metal salt is usually made of a
rubidium salt. Organic compounds containing nitrogen or phosphorus react with alkali
metals during combustion under the formation of anions such as cyanide and phosphoruscontaining anions as well as electrons. The response of nitrogen-containing organic
compounds can be up to 106 times higher than that of organic substances not containing
nitrogen. The lower limit of detection for nitrogen and phosphorus-containing organic
substances is about 10–10 g. The main application area for this detector is the determination
of drugs and metabolites in biological material.
14.9.3
Thermal Conductivity Detector
The thermal conductivity of a gas changes when the gas is mixed with other substances.
When the effluent from the GC column passes through a cell with a heated filament the
resistance of the filament changes with changes in the composition of the effluent and
creates a signal. The thermal conductivity detector (TDC; Figure 14.11) is therefore
universal and it is suitable for the determination of water vapor. However, the TCD is
less sensitive than the FID and is therefore less often used, but it is the detector of choice for
GC determination of water in pharmaceuticals.
14.9.4
Electron Capture Detector
The electron capture detector (ECD; Figure 14.12) is a selective detector suited for the
determination of compounds with a high affinity for electrons. The detector house contains a
radioactive foil typically with 63 Ni emitting b-electrons which by collision with the carrier
Gas Chromatography
205
Electrical leads to power
supply and external
bridge circuit
Gas
flow out
Gas
flow in
Filament
Figure 14.11 The thermal conductivity detector
gas molecules generate a plasma of electrons. Thus a constant current in the detector
provides a background signal which is changed when electron-capturing compounds enters
the detector. Compounds containing halogens or nitro groups have high affinities for
electrons and the lower limit of detection is at the femtogram level. The detector is applied in
pharmaceutical analysis and bioanalysis where analytes have been derivatized with
Insulator
-
+
Exhaust
Radioactive
source
From GC
column
Figure 14.12 The electron capture detector
206
Introduction to Pharmaceutical Chemical Analysis
polyhalogenated reagents. If the analytes do not contain halogens a number of derivatizing
reagents containing fluorine are available to make analytes electron capture active.
14.9.5
Mass Spectrometry Detector
A mass spectrometer (MS) is used both to quantify substances and to provide structural
information about the substances. In the mass spectrometer the analytes are ionized under
vacuum, usually by being bombarded by an electron beam. Normally the energy of the
electron is controlled at 70 eV. Ionized molecules are broken down into smaller pieces
called fragments. The ionized molecules and any fragments are first separated by their
mass/charge ratio (m/z) and then detected by a detector. A mass spectrum is a plot of the
relative intensity of ions formed from a given compound as a function of their m/z value.
The mass spectrum is very characteristic for a given compound and is said to be the
fingerprint of the compound.
Therefore, GC-MS is a standard method for drug analysis and forensic science.
Identifications are based upon a comparison of retention data from the GC separation
and from the mass spectra of the substances. MS is discussed in more detail in
Chapter 16.
14.10 Derivatization
Many drug substances contain polar functional groups such as hydroxyl groups, phenolic
groups, amino groups or carboxylic acid groups. These groups lower the volatility of the
substances due to intermolecular forces like hydrogen bonds or ionic interactions. Furthermore, the polar substances often interact with surfaces by adsorption, resulting in bad
chromatographic properties like peak tailing. Problems with low volatility and bad
chromatographic properties can be eliminated by derivatization. The purpose of derivatization is to create volatile and thermally stable derivatives of polar substances to improve
GC behavior and also to improve detectability. The derivatization blocks the polar intra- and
intermolecular forces, thus enhancing volatility.
14.10.1
Silylation
By silylation active H atoms are replaced by a trimetylsilyl group as shown in Figure 14.13.
Chemical groups such as OH, COOH, NH2 and NH are well suited for silylation.
Derivatives are thermally stable, volatile and suitable for GC. A large number of silylating
reagents are available from different suppliers.
CH3
CH3
R OH + Cl
Si
CH3
CH3
R O
Si
CH3
+ HCl
CH3
Figure 14.13 Silylation of a hydroxyl group
Gas Chromatography
CH2 O COR
HC O COR
KOH
R COOCH 3
CH2 OH
CH3 OH
+
HC OH
R COOCH 3
CH2 OH
CH2 O COR
207
R COOCH 3
Figure 14.14 Methylation of fatty acids in oils
14.10.2
Alkylation
Alkylation is a common term for derivatization where an active hydrogen is replaced by an
alkyl group. An example is the determination of fatty acids in oils according to the
pharmacopoeias. Fatty acids exist as triglycerides in oils. The fatty acids are released
by hydrolysis and separated by GC as methyl esters. By using methanol and potassium
hydroxide the reaction is done in one step as shown in Figure 14.14.
14.10.3
Acylation
Acylation replaces an active hydrogen atom with an acyl group. Both anhydrides and acid
chlorides are used as reagents, and groups such as OH, NH2 and NH can be derivatized.
Reagents usually contain fluorine atoms because derivatization reagents containing fluorine
atoms provide stable volatile derivatives suitable for GC. Commonly used reagents are
trifluoracetic acid anhydride or anhydrides of pentafluorpropionic acid or heptafluorbutyric
acid. Amino acids are nonvolatile due to the strong intermolecular ionic interactions.
Figure 14.15 shows the derivatization of an amino acid involving acylation and alkylation.
The derivative is volatile and also suitable for electro capture detection.
H 2N
+ H3N
COOH
C
COOC
α-amino acid
R
R
O
O
H 2N
COOH
+
C
F3 C
C
C
Acylation
CF3
HN
R
+ CH3OH
Η
+
+ CF3COOH
R
O
COOH
C
COOH
C
O
O
CF3
HN
O
F3 C
R
C
C
C
CF3
HN
COOCH3
C
Methylation
+ H2O
R
Figure 14.15 Derivatization of an a-amino acid with trifluoracetic acid anhydride and
methanol
208
Introduction to Pharmaceutical Chemical Analysis
14.11 The Uses of GC
In a pharmaceutical context GC is used for the determination of volatile impurities in drug
substances and preparations. These are primarily residual solvents coming from production,
but GC is also the official method for determination of the composition of essential oils and
the fatty acids in oils. Essential oils are added as excipients to pharmaceutical products
because of their odor and taste characteristics. Essential oils contain a complex mixture of
readily volatile components and capillary GC is therefore the method of choice. Vegetable
fats and oils are used as excipients in ointments and creams. GC is used for the analysis of the
composition of fatty acids in oils and to determine the content of sterols.
The determination of residual solvents in raw materials and drug preparations is
performed as prescribed in Ph.Eur. and USP using an injection technique called headspace
injection. The sample is pulverized and suspended in water in a test glass as shown in
Figure 14.16.
The headspace vial is heated in a water bath and the solvent evaporates to the space above
the sample solution – the headspace. The sample vial is closed with a membrane that
prevents the solvents from disappearing from the vial. A sample of 0.5–2.0 ml of the
headspace is collected with a gas-tight syringe and the sample is injected into the GC.
GC is also used to analyze drug substances and their metabolites in biological materials.
A GC analysis can separate substances with large differences in volatility. Because of this,
Step:
1
2
Sampling
Gas tight
syringe
3
4
Injection
into GC
Septum
Head
Space
Sample
Analytical
GC column
1 ml/min
Figure 14.16 Headspace injection of residual solvents
Gas Chromatography
209
GC is suitable for screening biological materials such as blood or urine for unknown
compounds, which is needed in toxicological and doping analysis where the purpose of the
analysis is to identify known or unknown substances that have been ingested. For such
analysis the use of combined gas chromatography–mass spectrometry (GC-MS) or LC-MS
are preferred. The substances separated by gas chromatography are carried on to the
mass spectrometer where the mass spectra of the substances are detected. The data
(retention time) from the GC separation and the data from the mass spectrum are used
for identification.
GC is one of the standard methods for the analysis of pollutants and is an important
analytical method in the food industry, perfume industry and the oil industry.
14.12 More Advanced GC Techniques
This chapter gives a basic introduction to gas chromatography. GC was invented as far back
as 1952 and became a very important analytical technique before HPLC was invented.
Progress continues to be rapid in GC and examples of new developments are, for example,
comprehensive two-dimensional GC (2-D GC), fast GC analysis where highly efficient GC
analysis is performed in few minutes using very steep temperature programming, detection
by atomic emission and time of flight mass spectrometry, and in applications to process
analytical chemistry.
14.13 Summary
Gas chromatography is the method of choice for the analysis of volatile substances.
The areas of application cover contaminants in pharmaceutical raw materials and products
such as solvent residues, analysis of essential oils, fatty acids and sterols in plant material.
GC is also used to analyze drug substances in biological material.
15
Capillary Electrophoresis
In its basic set-up electrophoresis is a separation technique based on the migration of
charged species/ions in an electrical field. Thus it is not based on distribution between two
immiscible phases and it is therefore not a chromatographic technique. However, the system
can be modified to become a combination of electrophoresis and chromatography. The
driving force in electrophoresis is a voltage gradient applied between two electrodes in a
buffer system. Charged analytes introduced into the system migrate towards the electrode
with opposite charge. In the present text only electrophoresis in capillaries and related
techniques are presented.
15.1 Principle and Theory
Charged compounds in a solution begin to migrate when an electric field is imposed on the
solution as shown in Figure 15.1. Positively charged ions (cations) migrate towards the
negative electrode (the cathode) as a result of electrostatic attraction, while the negatively
charged ions (anions) migrate towards the positive electrode (the anode).
The velocity of an ion obtained in a given system can be given by:
n ¼ me E
ð15:1Þ
where n is the ion velocity, me is the electrophoretic mobility of the ion and E is the applied
electrical field.
The electrical field is also expressed as volts per centimeter (V/L). When an ion travels
through a solution it is exposed to a frictional force, and when Stoke’s law is applied me can
be expressed as:
z
ð15:2Þ
me ¼
6prh
where z is the charge of the ion, r is its radius and h is the viscosity of the solvent. From the two
equations it can be concluded that the mobilitity of the ion increases with increasing charge
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
212
Introduction to Pharmaceutical Chemical Analysis
+
−
Voltage
+
+
−
−
Figure 15.1 Principle of electrophoresis: ionized molecules move in solution when an electric
field is imposed
but slows down with increasing radius and increasing solvent viscosity. An increase involtage
provides faster separations. In a given capillary electrophoresis (CE) system the charge and
the size of the ion is decisive for the electrophoretic mobility of the ion. Thus small ions travel
faster than larger ions and multiple charged ions travel faster than singly charged ions.
When operating with small molecules with a mass less than 103 the size of the ions is
considered proportional to the molecular mass, but when it comes to larger molecules like
proteins the actual size of the analyte ion is of importance and depends on how the threedimensional structure of the proteins is expressed in the actual buffer solution.
A prerequisite for electrophoresis is that the analytes are charged. Thus pH plays an
important role in the separation and a buffer is therefore used as the conducting liquid
(running buffer) in the capillary. Ions of similar size (mass) but with a small difference in
their pKa values can be separated if careful control of the pH is applied. At a pH value
between the two pKa values of the two analytes separation can be achieved due to the
difference in degree of ionization (Figure 15.2).
µ
Full charge
µ1
µ2
Zero charge
2
pKa1 pKa2
12
pH
Figure 15.2 Electrophoretic mobility (m1 and m2) of two weak acids as a function of pH in a
running buffer
Capillary Electrophoresis
213
15.2 Electroosmotic Flow
Capillary electrophoresis is most often performed in narrow silica capillaries typically with
an internal diameter of 50–75 mm. Other materials can be used, for example, Pyrex glass or
Teflon, but fused silica capillaries are preferred because of the robustness achieved when
coated on the outside with a polyimide layer. Common for these materials is that, in contact
with an aqueous solution, they achieve a negative charge on the inner surface. In the case of
silica the silanol groups have a major impact on the surface charge.
Silanol groups are acidic groups, and they deprotonate when the pH is higher than about 4.
The average pKa value of a silica surface is about 6–7. This means that the negative charge
on the inner surface of the silica capillary increases in the range from about 4 to 9. The
negative charges, which are fixed on the capillary surface, are balanced by cations in the
separation buffer to create a charge balance. In this way a double layer is built at the surface,
creating a potential difference reaching from the capillary wall into the buffer for a certain
distance. The potential is known as the zeta potential. The cations close to the inner surface,
which typically can be sodium ions, are hydrated, which means that they are bound to the
surrounding water molecules, which in turn are bound to other water molecules via
hydrogen bonds. When a potential is applied across the capillary the solvated cations
migrate towards the cathode and in this way drag the bulk solution in the capillary in the
same direction. This is called electroosmosis.
The generated electroosmotic flow (EOF) is strongly dependant on the pH. At pH > 7 a
relatively high flow is obtained (in the range of 10 cm/min) but even at low pH in 0.01 M
phosphoric acid an EOF is observed (0.5–1.0 cm/min). The EOF decreases with increasing
buffer strength as the increase in buffer ions suppresses the zeta potential and thus the
surplus of cations close to the surface. The reduced surplus of cations at the surface results in
a reduced drag on the bulk solution.
The bulk flow (EOF) in the capillary is generated at the same time at any part of the inner
surface of the capillary and therefore results in a so-called plug flow. This should be seen in
comparison to HPLC where a parabolic flow is generated due to the pumping from one end
of the system. Radial dispersion takes place across the parabolic flow profile and this type of
dispersion is eliminated in CE (Figure 15.3). Peak broadening effects due to mass transfer
between phases is also eliminated in CE. Therefore a high number of theoretical plates can
be generated in a fairly short capillary.
Flow profiles
ν0
Ideal plug flow
ν0
ν0
Electroosmotic flow
Parabolic flow
Figure 15.3 Flow profiles: a theoretical plug flow, an electroosmotic flow and a parabolic flow
214
Introduction to Pharmaceutical Chemical Analysis
The two primary causes of dispersion in CE are the longitudinal diffusion and electrodispersion. Diffusion of the analyte molecules is only of importance in the longitudinal
direction of the capillary, and the diffusion is low because it is in the liquid phase.
Electrodispersion is caused by a difference in mobility between the analyte ions and the
ions of the running buffer. A lower mobility of analyte ions compared to buffer ions results in
tailing peaks and a higher mobility of analyte ions results in fronting peaks. This is the case
when concentrations are about equal in the two zones as low mobility is a sign of low
conductance and thus a higher electrical field and visa versa.
The EOF in CE is thus in many ways comparable to the flow generated by a pump in
HPLC but the flow profile is different. CE performed in free solution using a simple buffer is
also called capillary zone electrophoresis (CZE).
15.3 Instrumentation
In a fairly simple CE (Figure 15.4) the separation takes place in a small capillary typically of
fused silica with an inner diameter of typically 50 or 75 mm filled with an aqueous buffer.
The driving force for separation is a high voltage power supply that can supply voltages
between 0 and 30 kV. A constant voltage between the electrodes is important for achieving
repeatable analysis. Some instruments are also capable of running at constant current
or constant power (watts). The voltage can be varied freely, but usually it is in the
range 10–30 kV, and high voltages provide rapid analysis, while low voltages give longer
analysis times.
The capillary of a length of typically 30–70 cm is placed between two vials containing the
running buffer. An electrode connected to the power supply is placed in each buffer vial. After
a few nanoliters of sample are introduced into one end of the capillary (see Section 15.5) a
voltage of typically 10–30 kV is applied to the system across the two electrodes.
The voltage applied initiates the EOF in the system as well as the electrophoretic
migration of the ionized analytes according to their mass or size to charge. Neutral analytes
migrate with EOF.
+
Power supply
-
Fused-silica capillary
Detector
Buffer
Buffer
Figure 15.4 Diagram for the CE instrument
Capillary Electrophoresis
215
The observed mobility (apparent mobility, mapp) of an analyte is a function of the
electrophoretic mep and the electroosmotic mobility meo:
mapp ¼ mep þ meo
ð15:3Þ
The apparent mobility can also be written as:
Ld
t
mapp ¼ m
V
Lt
ð15:4Þ
where Ld is the length of the capillary from injection to the detector, Lt is the total length of
the capillary, V is the voltage and tm is the migration time of an ion for distance Ld.
The number of theoretical plates obtained in a CE separation is calculated by the formula:
N¼
mapp V Ld
2D Lt
ð15:5Þ
where D is the diffusion coefficient. Typical diffusion coefficients are given in Table 15.1.
In the basic set-up of CE, samples are introduced at the anodic end of the capillary. The
separated analytes are then detected near the cathodic end of the capillary. Running at pH 7
EOF is sufficiently high to be able to also carry the anions to the detection site although the
anions have an electrophoretic migration towards the anode. The substances in the sample
injected at one end of the capillary are separated and arrive at the other end at different times.
Typically a UV detector is used for on-capillary detection. From a UV source UV light is
transmitted across the capillary. The absorption of UV light performed by the analytes passing
the detection window is transformed into a detector signal which is plotted as a function of
time. This plot is called an electropherogram (Figure 15.5) and shows a peak for each of the
substances having UV absorption and that were separated during the CE analysis. The UV
detectors used are often based on diode arrays and thus full UV spectra of analytes can be
recorded during analysis. An electropherogram is analogous to a chromatogram, but in the CE
the term electropherogram is used to highlight that this is not a chromatographic separation.
Using UV absorption spectrophotometry for the on-capillary detection of analytes
provides poor sensitivity compared to HPLC because the effective light path is equivalent
to the small internal diameter of the capillary. One possibility to improve this is to widen the
internal diameter at the detection window to form a so-called “bubble cell.” This can
Table 15.1 Selected diffusion coefficients at 25 C
Solute
Solvent
Diffusion coefficient (m2/s)
H2O
CH3OH
Hþ
OH
Kþ
Na þ
Serum albumin
H2O
H2O
H2O
H2O
H2O
H2O
H2O
2.3 109
1.6 109
9.3 109
5.3 109
2.0 109
1.3 109
0.059 109
216
Introduction to Pharmaceutical Chemical Analysis
AU
COOH
NH2
H
N
COOCH3
O
HOOC
N
H
N
NH2
0.00
2.00
4.00
6.00
8.00
10.00 min
Figure 15.5 Electropherogram
improve the signal to noise by a factor of 3–5. Another possibility is to bend the capillary to
form a “Z” using the diagonal in Z as the detection cell improving the signal to noise by a
factor of about 10. The Z-cell, however, leads to some band broadening.
Other detection principles in CE are also possible but it is often much more selective and
thus fewer analytes can be detected. Using laser induced fluorescence as detection it is
possible to detect a few or even a single molecule, but this molecule is present in a very small
volume and thus the concentration sensitivity is less impressive. Table 15.2 gives a number
Table 15.2 Detectors for CE
Detection method
Mass LOD (moles)
Concentration
LOD (molar)
Remarks
UV-Vis
1013–1016
105–108
Fluorescence
1015–1017
107–109
Laser-induced
fluorescence
Amperometry
1018–1020
1014–1016
“Universal”, diode array
offers full spectra
Sensitive for selected
substances or derivatives
Very sensitive, expensive
1018–1019
1010–1011
Conductivity
1015–1016
107–108
Mass spectrometry
1016–1017
108–109
Indirect UV,
fluorescence
10–100 less sensitive than direct method
Sensitive but only for
electroactive compounds
Universal, requires special
equipment
Sensitive, gives structural
information, complicated
Universal, but less sensitive
Capillary Electrophoresis
Low field
217
High field
+ + ++
+ + + +
+ +
+ +
+
-
+
ep
eof
-
-
ep
+ + + ++ +
+ + +
+ + + +
+++
+ + ++
+
+ ++
+
+
Figure 15.6 Stacking of cations at low pH
of detection principles, showing the typical absolute detection limit as well as the
concentration detection limit for each detection principle.
Tuning of the sample conductivity relative to the conductivity of the running buffer can
lead to a concentration of the introduced sample – a concentration step that can improve
the signal to noise by a factor between 10 and 100. This principle is called sample
stacking (Figure 15.6). The stacking can be obtained when the buffer concentration in the
sample is one-tenth or less than that of the running buffer. The electric field is inversely
proportional to the conductivity and a low concentration in the sample results in a high
electrical field. The electrophoretic mobility of the sample ions is high in the sample zone.
When the ions in the sample zone reach the boundary to the surrounding running buffer
they experience the lower field in the running buffer and the migration rate decreases.
In this way the sample ions are concentrated at the boundary between the sample zone
and the running buffer. Subsequently the concentrated sample zone is separated in the
running buffer.
Considering the reverse situation where a high concentration of ions is present in the
sample, the mobility of ions is low until they reach the boundary to the surrounding running
buffer where the electric field is higher. The ions then speed up, resulting in band
broadening. Thus high salt concentrations in the sample ruin the CE separations.
15.4 The Capillary
Separations are carried out as previously described in fused silica capillaries with a diameter
of either 50 or 75 mm. The length of these is typically 30–70 cm. Long capillaries are used if
you want to separate a large number of substances. Shorter capillaries are used if you have a
simple sample to be analyzed in a short time.
218
Introduction to Pharmaceutical Chemical Analysis
The fused silica capillaries used for CE are coated on the outside with a thin layer of
polyimide polymer. This layer makes the capillaries extremely flexible and prevents
breakage. To be able to perform on-capillary UV detection the polyimide layer, which
absorbs UV light, has to be removed in order to form a detection window of a few
millimeters of length. The polyimide layer can be removed from a fused silica capillary by
simple burning (Figure 15.7). If the capillary is coated on the inside surface removal of the
polyimide layer is performed using concentrated sulfuric acid. This is done at a suitable
distance from the outlet end of the capillary. When the polyimide layer has been removed the
capillary becomes very fragile at the detection window.
Electrophoresis is a temperature-dependent process, where the migration rate of
analytes through the capillary varies with temperature. This is important to consider as
the current generated, when voltage is applied, generates heat like in an electrical bulb. An
increase in current gives an increase in heat (Joule heating) and, if the temperature
becomes too high in the capillary, air bubbles may be formed, resulting in loss of
conductivity. This makes the CE stop. The temperature rise is also the reason why
capillaries with relatively small inner diameters are used. An increase in inner diameter, all
other things being equal, results in an even higher increase in current and thus in
temperature. Control of temperature around the capillary is therefore essential to be
able to obtain repeatable and reproducible separations and peak areas. However, although
the temperature around the capillary is controlled to a certain temperature it is most
likely that the temperature inside the capillary is higher as the temperature equalization
between the inside and the outside of the capillary takes some time, but at least the inside
temperature is kept at a steady state.
When the temperature rises in the capillary the viscosity of the running buffer decreases
and EOF increases. Also the electrophoretic migration of ions changes as the frictional
forces are reduced.
Before the capillary is taken into use it is flushed with 1 M sodium hydroxide, 0.1 M
sodium hydroxide and finally with the running buffer. In this way a reproducible amount of
silanols on the inner surface is generated. From time to time it may be necessary to repeat
this rinsing procedure in order to obtain repeatable migration times.
15.5 Sample Introduction
In order to achieve good separations, it is important that the sample is introduced in a small
volume. A 50-cm capillary with 50 mm internal diameter holds just about 1 ml (1000 nl). The
introduced volume should not exceed about 1% of the capillary volume. Thus only few
nanoliters are to be introduced. It is obvious that it is difficult to inject such a small sample
volume repeatably using a syringe. Therefore two other techniques are used for sample
introduction. Most commonly used is hydrodynamic injection (Figure 15.8).
The inlet end of the capillary is placed in the sample vial and a well defined pressure is
placed on the vial for a well defined time span (seconds). The capillary end is then moved
from the sample vial back to the buffer vial and the voltage is applied to start the separation.
Instead of using an overpressure on the inlet vial a vacuum in the outlet can also be used.
Another hydrodynamic principle is siphoning where the sample vial with the inlet end of the
Figure 15.7 Detection window on the capillary and its preparation
Fused silica capillary
(polyimide coated)
ca 5 mm
Teflon sleeves
Capillary Electrophoresis
219
220
Introduction to Pharmaceutical Chemical Analysis
Hydrodynamic
Vacuum
Pressure
Sample
Siphoning
Sample
Sample
Electro kinetic
Sample
+
Figure 15.8 Sample injection and subsequent separation
capillary inserted is raised to a higher level than the outlet vial. Performed in a controlled
manner this principle results in a well defined sample introduction. The volume injected in
hydrodynamic mode can be calculated from the formula:
Volume ¼
DPpd4 t
128hLt
ð15:6Þ
where DP is the pressure applied, d is the internal diameter of the capillary, t is the time
the pressure is applied, h is the sample viscosity, and Lt is the total length of the
capillary.
Electromigration is another technique used for sample introduction also called electrokinetic injection. In this technique a voltage is applied shortly over the capillary while
its inlet end is placed in the sample vial. The EOF generated brings a volume of sample
into the capillary while also the voltage influences the introduction of sample ions to
the capillary. The inlet end is subsequently moved back to the buffer vial and the voltage
is applied.
The hydrodynamic principle results in the introduction of a representative volume of the
sample to the capillary, while the electromigration discriminates one type of ions against
another.
Most CE instruments are fully automated and computer-controlled systems. The
instruments are equipped with an autosampler rack where you can set up a larger number
of samples and analyze them automatically.
Capillary Electrophoresis
221
15.6 Capillary Zone Electrophoresis; an Example
In capillary zone electrophoresis (CZE) a simple aqueous buffer is used as the running
buffer. In this example a mixture of phenylethylamine, codeine, phenobarbital and vanillic
acid is to be separated by CZE. A 0.01 M phosphate buffer at pH 7.2 is selected for the
running buffer. This generates a sufficiently high EOF to be able to detect the anionic
analytes. A voltage of þ 20 kV is applied and an electropherogram as shown in Figure 15.9
4
Mr
pKa
121.2
9.8
299.4
8.2
2
NH2
3
1. Phenethylamine
H 3 CO
O
NCH 3
HO
2. Codeine
H
N
O
O
H 3C
NH
1
232.2
7.3 ; 11.5
168.2
4.5 ; 7.5
O
3. Phenobarbital
COOH
OCH 3
OH
4. Vanillic acid
0
3
min
Figure 15.9 Electropherogram of a mixture of phenylethylamine, codeine, phenobarbital and
vanillic acid
222
Introduction to Pharmaceutical Chemical Analysis
is obtained. From the molecular structures, molecular weights (Mr) and the pKa values it is
possible to assign the peaks in the electropherogram. In principle it is also possible to predict
the migration order of the analytes before performing the electrophoresis.
Phenethylamine is a small molecule which is fully ionized at pH 7.2 due to a pKa of 9.8.
Codeine is a somewhat larger molecule and with a pKa of 8.2 it is only partly ionized (90%)
at pH 7.2. Both these two analytes have an electrophoretic migration towards the cathode
and the detector and migrate in front of the EOF and therefore are detected first.
Phenylethylamine reaches the detector first. Phenobarbital and vanillic acid are both acidic
in nature. Only one of the protons on phenobarbital is partly ionized (ca. 50%) at pH 7.2 as
the pKa value is 7.3. In vanillic acid the carboxylic acid group (pKa 4.5) is fully ionized at pH
7.2 and the phenolic group (pKa 7.5) is partly ionized at this pH (a little less than 50%).
Vanillic acid, being the smaller of the two anionic analytes and having about 1.5 negative
charge, has the highest electrophoretic migration rate of the two. But the electrophoretic
migration is towards the anode away from the detector and the vanillic acid is therefore
detected as the last one of the four analytes.
This example show how knowledge about molecular structure and the Mr and pKa values
are important parameters to handle when designing and predicting a separation in CZE.
15.7 Micellar Electrokinetic Chromatography
Capillary zone electrophoresis (CZE) as described above is very suitable for the separation
of acidic or basic drug substances and other ionizable solutes while all neutral substances
migrate together with EOF. However, in many cases it is an advantage if ionizable as well as
neutral substances can be separated simultaneously. When using the so-called micellar
electrokinetic chromatography (MEKC) such separations can be performed (Figure 15.10).
MEKC is very similar to CZE, the difference is that a surfactant has been added to the
running buffer. Above a certain concentration the surfactant molecules form micelles in the
Figure 15.10 Schematic representation of MEKC and a single micelle
Capillary Electrophoresis
223
running buffer. Sodium dodecylsulfate (SDS) is the most often used surfactant for this
purpose. The globular SDS micelles are an aggregation of dodecylsulfate ions forming a
lipophilic core in which the hydrophobic hydrocarbon chains are oriented towards the core
of micelles, while the negatively charged sulfate groups are facing the water phase.
The SDS micelles have an electrophoretic migration towards the anode away from the
detector. Therefore it is necessary to have an EOF sufficiently high to bring all analytes, also
those with a distribution to the micelles, to the detector. This is achieved using a sufficiently
high pH of the running buffer – typically above pH 7. Neutral substances migrate with EOF
when they are outside the micelles but when they are distributed to the micelles they migrate
together with the micelles towards the anode. Solutes with a different partition to the
micelles migrate with a different rate towards the anode and thus are separated. Polar
analytes are detected before the more hydrophobic ones which have a larger distribution to
the micelles. The micellar phase is called a pseudostationary phase and the separation
mechanism in a MEKC system is a mixture of electrophoresis and chromatography
(partition between phases).
Neutral substances can thus be separated in MEKC according to their hydrophobicity. An
example of how different neutral compounds migrate is shown in Figure 15.11.
The separation of eight different steroids is performed with very high efficiency and
according to their hydrophobicity. The small peak after approximately 9 min is a signal
originating from the solvent of the injected sample, and as it has little affinity to the micelles
it can be used as a measure of the EOF. The steroids have significantly longer migration
times than the EOF, which shows that they exhibit relatively strong interactions with the
micelles, and this is natural in light of the generally hydrophobic nature of steroids.
6
mAU
2
3
7
4
8
5
1
0.02
0.01
8
12
16
20
min
Figure 15.11 Separation of eight steroids using a borate buffer (pH 8.45) containing 100 mM
sodium cholate (micelle) as separation buffer: (1) triamcinolon, (2) hydrocortisone, (3) betamethasone, (4) hydrocortisonacetate, (5) dexamethasone, (6) triamcinolonacetonide, (7) fluocinolonacetonide, (8) fluocinolon
224
Introduction to Pharmaceutical Chemical Analysis
Figure 15.12 MEEKC of opium alkaloids. Capillary: 32 cm 25 mm from short end. Voltage:
30 kV. Running buffer: 1-chloro-pentane þ lithium dodecylsulfate þ n-butanol þ 10 mM
sodium tetraborate pH 9.2 (0.8 : 3.3 : 6.6 : 89.3, v/w/v/v)
Microemulsion electrokinetic chromatography (MEEKC) is a technique very similar to
MEKC. In MEEKC a nonmiscible organic solvent (e.g., n-heptane) is added to the buffer
containing the surfactant (e.g., SDS) and a co-solvent is also added (e.g., n-butanol). The
internal parts of the micelles are now loaded with the organic solvent. Figure 15.12 show the
fast separation of opium alkaloids. The injection is made from the end of the capillary close
to the detection window and using a negative voltage in order to reverse the EOF. This
delivers the separated molecules to the detector in a very short time.
Capillary electrochromatography (CEC) is another way to combine electrophoresis and
chromatography. In this technique the electrophoresis is performed in capillaries packed with
a stationary phase similar to what is known from HPLC. Silica-based particles chemically
modified with octadecyl groups and with a particle size of 3 or 5 mm are often used for CEC.
15.8 Chiral Separations
A large number of drug substances are racemic mixtures or optically active enantiomers and
in biological systems enantiomers can exhibit different pharmacological effects. Therefore
there is a great need to be able to separate such chiral compounds. In CZE and MEKC two
enantiomers of a given substance have the same physicochemical characteristics and show
the exact same time of migration and consequently cannot be separated. To achieve chiral
separations in CE two overall principles can be used: the indirect or the direct. The indirect
Capillary Electrophoresis
225
principle is based on derivatization with a chiral reagent to form diastereomers. Diastereomers have different physicochemical characteristics and may therefore be separated using
MEKC. In the direct mode the native chiral molecules are separated and a chiral selector has
to be incorporated into the CE system either as a stationary coating of the internal surface of
the capillary or by adding a chiral reagent (selector) to the running buffer. Chiral selectors
can be used in both CZE and MEKC. Chiral selectors are usually relatively large molecules
that have different ability to form complexes with enantiomers of a given substance.
Because the spacious structure of the two enantiomers is different, they may form
complexes of different strength with the chiral selector. When bound to the chiral selector
an enantiomer exhibits a different migration rate compared to the rate as a free molecule.
This leads to the separation of the two enantiomers.
Typically, different forms of cyclodextrins are used as chiral selectors in CE. Cyclodextrins are a group of cyclic oligosaccharides consisting of six, seven or eight glucose units and
are referred to, respectively, as a-, b- or g-cyclodextrins. A huge number of cyclodextrins
modified by chemical synthesis are available both in charged and uncharged forms. Chiral
separations can be optimized by using different cyclodextrins or even by mixtures of
different cyclodextrins. Figure 15.13 shows a structure of b-cyclodextrin.
In principle any chiral compound can be used as a chiral selector in CE, but unfortunately
it is very much a case of “trial and error” to find the best selector for a given chiral separation.
Figure 15.14 shows an example of a simultaneous chiral separation of five enantiomeric
pairs (racemates) of anaesthetic drug substances using dimethyl-b-cyclodextrin as chiral
selector. The resolution between the enantiomers in each substance in general increases with
the time the substances spend in contact with the chiral selector.
Besides cyclodextrins, proteins, polysaccharides, optically active surfactants, antibiotics
and so on have been used as chiral selectors.
15.9 Coated Capillaries
Capillaries can be coated on the inside surface due to different reasons:
1. A simple coating with an organic polymer (e.g., polyvinylalcohol) more or less
eliminates EOF as the silanol groups are masked. At the same time adsorption of
proteins onto the surface is prevented. These coated capillaries are often preferred for the
analysis of proteins.
2. The inner surface can also be coated with a protein acting as a chiral selector.
Enantiomers can then be separated when they have different affinities to the stationary
protein phase.
3. The purpose of a coating can also be to reverse the EOF. If a quaternary ammonium
compound is added to the running buffer it adsorbs through strong ionic interactions onto
the inner surface and creates a surplus of positive charges at the surface instead of the
negative charges from the silanol groups. This results in the formation of a diffuse layer
of negative charges which, when voltage is applied, migrate towards the anode and drag
the water phase in this direction. Thus the EOF is reversed. If the polarity of the voltage
subsequently is reversed the EOF is again directed towards the detector. In this mode the
anions reach the detector before the EOF marker and are then followed by the cations.
HO
O
O
O
HO
O
HO
OH
OH
HO
O
HO
OH
HO
O
HO
OH
OH
O
OH
OH
HO
HO
O
HO
O
O
OH
O
O
OH
OH
HO
HO
HO
HO
OH
OH
OH
OH
OH
OH OH
OH
OH
OH
OH
OH
OH
OH
OH
OH
Figure 15.13 Structure of b-cyclodextrin with the polar rim and the hydrophobic cavity
O
OH
O
Hydrophobic cavity
Polar rim
226
Introduction to Pharmaceutical Chemical Analysis
Capillary Electrophoresis
227
mAU
4
8
4
3R
1 1
6
2 2
3S
5
5
4
2
22
24
26
28
30
32
34
min
Figure 15.14 Chiral separation of enantiomeric anaesthetics with a phosphate buffer (pH 3.0)
containing 13 mg/ml dimethyl-b-cyclodextrin as chiral selector: (1) R- and S-mepivakaine, (2) Rand S-ethylpipecoloxylidide, (3) R- and S-ropivakaine, (4) R- and S-bupivakaine, (5) R- and
S-pentyl-pipecoloxylidide
4. In order to obtain a high EOF at a low pH the inner surface of the capillary can be coated
first with a quaternary ammonium polymer followed by an anionic polymer (e.g.,
polyvinylsulfonate; Figure 15.15).
As the large polymers have a very high affinity to the surface this kind of coating is nearly
permanent. The polyvinylsulfonate creates a high density of negative charges at the surface
OH
O
+N
Si
Si
O
Si
OH O-
O
Si
O
Si
O
Si
OH OH O+N
O-
O
O
Si
O
Si
O
O
Si
OH OH O-
+N
+
N
OH
Si
N+
O
Si
O
Si
OH OH
O
+N
Si
Si
O-
O Fused silica
SO3 - SO3 - SO3 - SO3 - SO3 - SO3 - SO3 - SO 3 - SO3 - SO3 - SO 3 - SO3 - SO3 - SO3 - SO3 -
SO3 - SO3 - SO3 - SO3 - SO3 - SO3 - SO3 - SO 3 - SO3 - SO3 - SO 3 - SO3 - SO3 - SO3 - SO3 -
+N
Hexadimethrine layer
Polyvinylsulfonate layer
Polyvinylsulfonate layer
+N
Hexadimethrine layer
+N
-N
OH OH O + OH
Si
Si
Si
Si Si Si Si Si Si Si Si Si Si Si
Fused silica
O
O
O
O
O
O
O
O
O
O
O
O
O
O
N
N+
O-+ OH OH
O-
OH OH O-
OH OH O-
Figure 15.15 Schematic representation of the coating for high EOF in the pH range 2–11
228
Introduction to Pharmaceutical Chemical Analysis
N
N
NH2
N
N
H
1. Didesmethylimipramine
N
2. Desmethylimipramine
3. Imipramine
N
N
N
+
N
O
4. Methylimipraminium
5. Imipramine-N-oxide
1+2+3
3
3 2
4
2
1 5
1
5
4
4
5
10
20
20
40
4
5
Minutes
Figure 15.16 Comparison of CZE, MEKC and non-aqueous CE for the separation of five closely
related analytes. Running buffer: left ¼ 0.05 M 6-aminocaproic acid (pH 4.0), middle ¼ 0.05 M
6-aminocaproic acid (pH 4.0) with 25 mM MAPS and 15 mM Tween 20 added, right ¼ 25 mM
ammonium acetate, 1 M acetic acid in acetonitrile
Capillary Electrophoresis
229
and as sulfonic acids have a pKa of 0–1 a high number of negative charges are also present at
low pH.
15.10 Non-Aqueous CE
pH is only defined in dilute aqueous solution. When organic solvents are added it is still
possible to perform a pH measurement but the value now has to be assigned as pH showing
that it is not a true pH value. When using neat organic solvent a pH value may still be
measured as long as there is conductivity in the system. Electrophoresis can therefore also be
performed in polar organic solvents like methanol or acetonitrile. When electrolytes and
analytes are in a non-aqueous environment they behave very differently than in an aqueous
solution. Water has a levelling effect on the ionization and in practice this is shown when
strong acids (like HCl, H2SO4, HNO3) are dissolved in water. They all show a similar acid
strength in water. When such acids are dissolved in an organic solvent they show very large
differences in their acid strength due to major differences in ionization. Such differences
may also be expressed with the analytes and therefore an improved separation of closely
related substances (Figure 15.16) can be expected in non-aqueous CE.
15.11 Summary
CE is a highly efficient separation method that produces significantly higher separation
efficiency than obtained in HPLC. It is not uncommon that the theoretical plate number in
CE is in the range 50 000 to 500 000, which in practice means that peaks in a electropherogram are very narrow. The high separation ability in CE can be used in two different ways. In
cases where you have relatively simple samples (typical in quality control of pharmaceutical
raw materials and preparations) separations can be performed in a very short time. In cases
of complex samples CE can be used to effectively separate/analyze a large number of
different components. In comparison with HPLC, where significant amounts of organic
solvents are used in the mobile phase, CE is a very cost-effective and environmentally
friendly technology.
CE is a very attractive separation technique and is complementary to HPLC. Especially
when analyzing larger molecules, which have a slow distribution between phases in HPLC
resulting in broad peaks, CE can provide very efficient separations. However, because the
detection normally takes place directly on the capillary by UV spectrophotometry, the
concentration sensitivity of CE is limited (see Table 15.2).
CE is an official technique of pharmaceutical analysis in the pharmacopoeias and it is
used for quantitative analysis relating to pharmaceutical products and preparations.
16
Mass Spectrometry
This chapter discusses the principles of identification and quantitative analysis based on
mass spectrometry (MS). MS is an official technique in the European Pharmacopoeia
(Ph.Eur.). MS is mainly used for the analysis of drug substances in biological samples.
Various MS techniques will be reviewed, and the most widely used instrumentation will
be discussed. Applications of MS are presented in Chapter 23 in connection with the
bioanalysis of drug substances.
16.1 Introduction
Mass spectrometry is performed using a relatively expensive and sophisticated instrument
called a mass spectrometer (MS). The MS is normally coupled to either gas chromatography
(GC-MS) or liquid chromatography (LC-MS). In the case of GC-MS, the analytes (drug
substances and other organic compounds) are bombarded with electrons in a vacuum inside
the mass spectrometer. Initially, this leads to some of the molecules being ionized either to
positive ions (molecules lose an electron) or negative ions (molecules take up an electron).
These ions are called molecular ions. The mass of an molecular ion is equivalent to the mass
of the original molecule because the weight of an electron that is either lost or taken up is
incredibly small. In some cases, molecular ions are unstable and very quickly split into
smaller fragments as chemical bonds in the molecules are broken. This process is called
fragmentation. Some of the fragments remain ionized (fragment ions), while others lose
their charge. Figure 16.1 shows an example of how ionization and fragmentation can take
place. When chlorambucil (antineoplastic drug) is bombarded with electrons in the mass
spectrometer, some of the molecules ionize to the positively charged molecular ions with
mass 303. A large proportion of these are unstable and will fragment. Fragmentation of
chlorambucil occurs due to several reactions, but only one of them is discussed at this stage.
The dominating process is the cleavage of CH2Cl from the molecular ion. The weight of
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
232
Introduction to Pharmaceutical Chemical Analysis
OH
Cl
O
N
+e
Cl
Cl
Mass = 303
Mass = 303
Cl
+
OH
O
N
+ 2e
O
N
Cl
OH
+
OH
O
N
Cl
+
+ CH2Cl
Cl
Mass = 303
Mass = 254
Figure 16.1 Ionization (upper reaction) and one major fragmentation (lower reaction) of
chlorambucil
CH2Cl is 49, and loss of this fragment from the molecular ion gives fragment ions with mass
303 49 ¼ 254.
Ionization and the subsequent fragmentation takes place in an ion source inside the MS.
The mass of the molecular ions and the molecular fragments that are still charged (fragment
ions) are then determined as the ions are accelerated out of the ion source and into the
mass analyzer. In the mass analyzer, ions are separated according to the ratio between the
mass (m) and the charge (z) in either a magnetic or electrostatic field (Figure 16.2).
The ratio between mass and charge is abbreviated m/z. The different ions are separated
based on their different deflection in the magnetic or electrostatic field inside the mass
analyzer. Ions with high m/z values are deflected to a lower extent than ions with a low m/z
value. In many cases, the ions carry a single charge, and in such cases the deflection is only
dependent on the mass of the ions. After separation in the mass analyzer, the ions are
transferred to the detector, which measures the amount or intensity of the different ions with
different m/z values.
Magnetic or electrostatic field
Mass analyzer
Ion source
Detector
Increasing m/z
Figure 16.2 Principle of mass spectrometry
Mass Spectrometry
233
Figure 16.3 Mass spectrum of chlorambucil
The result of a mass spectrometric analysis is plotted in a mass spectrum, where the
intensity (amount) of the ions is plotted in a bar graph as a function of m/z. For GC-MS of low
molecular drugs, the charge of the ions is normally þ1, and in practice the x-axis is a mass
scale. Figure 16.3 shows the mass spectrum of chlorambucil (which was featured in
Figure 16.1). A peak is observed at mass 303 from the molecular ion (Mþ) corresponding
to the molecular weight of the compound. In addition, the large peak of mass 254
corresponds to the loss of CH2Cl (M 49)þ, and the small peak at mass 268 corresponds
to the loss of Cl (M 35)þ.
What type of information can we get from a mass spectrum? Mass spectra can be used to
identify organic compounds because different substances have different mass spectra. An
unknown compound can thus be identified if the mass spectrum of the compound exactly
matches the mass spectrum of a known chemical reference substance. If this is not the case,
the compound can not be identified unambiguously, but mass spectrometry can still provide
important structural information because interpretation of mass spectra can provide
information about the following:
.
Molecular weight;
Elemental composition;
. Empirical formula;
. Functional groups.
.
This is discussed in more detail later. In addition, MS can be used for quantitative analysis.
This is also discussed at the end of this chapter.
16.2 Basic Theory
Atomic masses are measured in atomic mass units (amu), and 1 amu is defined as onetwelfth of the mass of the 12 C isotope. This means that the mass of the 12 C isotope is exactly
234
Introduction to Pharmaceutical Chemical Analysis
12.0000 amu. The mass of other atoms are measured relative to the weight of the 12 C
isotope. The 1 H isotope, for example, is 11.9068 times lighter than the 12 C isotope, and the
mass is therefore:
1
12:0000 amu ¼ 1:0078 amu
11:9068
ð16:1Þ
The 35 Cl isotope is 2.91407 times heavier than the 12 C isotope, and the mass is therefore:
2:91407 12:0000 amu ¼ 34:9688 amu
ð16:2Þ
The mass of a certain molecule is calculated by the sum of the masses of the individual
isotopes. The mass of chlorambucil 12 C14 1 H19 35 Cl2 14 N 16 O2 (which contains the isotopes
12
C , 1 H , 35 Cl , 14 N , and 16 O ) is therefore:
14 12:0000þ 19 1:0078þ 2 34:9688þ 14:0031þ 2 15:9949 ¼ 303:0787
ð16:3Þ
In MS, masses are measured in amu, or in Daltons which is the same as amu (1 amu ¼ 1
Dalton). Atomic and molecular masses as given above with three or four decimal places are
known exact masses. Mass spectrometry with high resolution measures exact masses,
whereas in MS with low resolution, the masses are rounded to the nearest integer. Masses
rounded to the nearest integer are called nominal masses. Measured in nominal masses, the
molecular mass of chlorambucil (12 C14 1 H19 35 Cl2 14 N 16 O2 ) is therefore:
14 12þ 19 1þ 2 35þ 14þ 2 16 ¼ 303
ð16:4Þ
As most routine mass spectrometry is performed with low resolution, we principally use
nominal masses in this chapter.
As mentioned in Section 16.1, mass spectra are bar plots where the quantities (intensities)
of the molecular ion and the fragment ions are plotted as function of the mass to charge ratio
(m/z value). In a mass spectrum, we can therefore see the fragment ions formed in the ion
source and in what relative quantities they are formed. In the magnified mass spectrum for
chlorambucil (Figure 16.4), for example, we see a large peak for mass 254. This tells us that
this particular fragment, corresponding to (M–CH2Clþ), is formed in large quantities.
In addition to this main fragment ion, also a large number of other ions occur, but these are
formed in substantially lower quantities (lower intensity). One example is at mass 268
corresponding to (M–Cl)þ. Most fragment ions from low molecular compounds contain
only one charge (z ¼ 1), and the m/z values are therefore practically equal to the mass of the
fragments (m/z ¼ m). Because of this, we can assume that the mass of the main ion for
chlorambucil is 254 and that the mass of the molecular ion is 303. In low-resolution mass
spectrometry, bars are plotted at nominal masses. This means that all exact masses, such as
recorded between 253.5000 and 254.5000 Dalton, are plotted as a single bar with mass 254.
The y-axis in a mass spectrum shows the amount or intensity of the different ions. A peak
or a bar with high intensity thus corresponds to the formation of many ions with that
particular mass. The intensity of the different ions will depend on the amount of sample that
enters the mass spectrometer; high amounts of sample in the ion source provide high
amounts of fragment ions. To make mass spectra independent of concentration, it is
common to scale the spectra relative to the most intense fragment ion. The most intense
fragment ion is called the base peak and the intensity of the base peak is scaled to 100%. The
other fragment ions are then scaled relative to the base peak.
Mass Spectrometry
235
254
256
303
305
307
304
Figure 16.4 Magnified mass spectrum of chlorambucil in the mass range 250–310
Unlike the other methods that have been discussed in this textbook, MS differentiates
between different isotopes, and this is important to take into consideration when interpreting
mass spectra. With chlorine as an example, there are two naturally occurring isotopes: one is
35
Cl isotope with mass 34.968855 (35) and the other isotope is 37 Cl having a mass of
36.965896 (37). The isotope with the lower mass is dominant and constitutes 75.77%,
while the heavier isotope constitutes 24.23%. Because these two isotopes have different
masses, the mass spectrum for chlorambucil (C14H19Cl2NO2), which contains two chlorine
atoms, has three peaks for the molecular ion (Figure 16.4): mass 303 corresponds to
molecular ions having two 35 Cl atoms, mass 305 contains one 35 Cl and one 37 Cl , while
mass 307 contains two 37 Cl atoms. Mass 303 is the highest signal as it is most probable that
both Cl atoms are 35 Cl . The probability for this is 0.750.75 ¼ 0.56. Mass 305 is lower in
intensity than mass 303 because it is less probable that one of the Cl atoms is 37 Cl . The
probability for this is 0.750.25 þ 0.250.75 ¼ 0.38, and the intensity of 305 is (0.38/0.56)
100% ¼ 68% of mass 303. Mass 307 is even lower because the probability for two 37 Cl
atoms in the same molecular ion is relatively low. The probability for this is
0.250.25 ¼ 0.06, and mass 307 corresponds to (0.06/0.56)100% ¼ 11% of mass 303.
It is not only chlorine that naturally occurs with multiple isotopes. In practice, all
elements frequently found in organic compounds exist as several isotopes except for
fluorine, iodine and phosphorus. Table 16.1 gives a list of the different isotopes and their
natural occurrence (%).
236
Introduction to Pharmaceutical Chemical Analysis
Table 16.1 Masses and occurrence of stable isotopes
Isotope
1
H
H
12
C
13
C
14
N
15
N
16
O
17
O
18
O
19
F
28
Si
29
Si
30
Si
31
P
32
S
33
S
34
S
36
S
35
Cl
37
Cl
79
Br
81
Br
127
I
2
Mass
Occurrence (%)
1.007825
2.014102
12.00000
13.003354
14.003074
15.000108
15.994915
16.999133
17.999160
18.998405
27.976927
28.976491
29.973761
30.973763
31.972074
32.971461
33.967865
35.967091
34.968855
36.965896
78.918348
80.916344
126.904352
99.985
0.015
98.90
1.10
99.634
0.366
99.762
0.038
0.200
100
92.23
4.67
3.10
100
95.02
0.75
4.21
0.02
75.77
24.231
50.69
49.31
100
As you can see in Table 16.1 carbon also exists naturally as two isotopes, namely the 12 C
isotope with mass 12.00000 (12) and the 13 C isotope with mass 13.003354 (13). The
12
C isotope dominates and constitutes 98.90%, while the 13 C isotope represents only
1.10%. In the molecular ion for chlorambucil, there are a total of 14 carbon atoms and
14 1.1% ¼ 15.4% of these molecular ions contain 13 C atom. So you can see a significant
peak at mass 304 (see Figure 16.4). For chlorambucil, the peak at mass 254 corresponds
to the loss of CH2Cl from the molecular ion (C14H19Cl2NO2 – CH2Cl ¼ C13H17ClNO2).
This fragment ion contains 13 carbon atoms and an average of 13 1.1% ¼ 14.3% of
the ions with the formula C13H17ClNO2 contains one 13 C atom. Therefore, we find a
significant peak at mass 255 (see Figure 16.4). As shown below, the isotope ratios are
important in connection with the interpretation of unknown spectra and can provide
significant structural information.
16.3 Electron Ionization
Mass spectrometry is frequently performed using electron ionization (EI) as the operating
principle in the ion source. Here, the substances to be investigated are bombarded by
electrons under vacuum conditions. Electron ionization is usually carried out with the MS
operated as a detector for gas chromatography (GC), where the substances after separation
in the GC instrument are transferred directly into the MS in the gas phase. GC coupled with
Mass Spectrometry
237
MS is abbreviated GC-MS, and this technique enables us to take a mass spectrum for each of
the compounds separated by the GC.
The EI ionization takes place under vacuum in the ion source, inside the MS, as the
analytes cross a beam of electrons obtained from a small wire (filament) of rhenium or
tungsten. The electrons released from the filament are accelerated in an electric field with a
potential of 70 V, which means that the electrons gain an energy of 70 eV. When the analytes
enters this electron beam, some of the molecules lose an electron as a consequence of
electrical repulsion (minus charges on the electrons repel each other):
Mþ e ! M.þ þ 2e
ð16:5Þ
This is an inefficient process; out of one million molecules, on average, only one molecule is
ionized. In spite of this, enough ions are normally produced in the ion source, and MS is thus
a very sensitive technique that can be used to detect very low concentrations.
Under electron ionization, M.þ is formed which is called the molecular ion. This has the
same mass as the original molecule, because the weight of the single electron that was
removed is negligible. The molecular ion is positively charged because one electron
was removed during the ionization. In addition to the charge, the symbol for the molecular
ions is usually supplied with a small dot to indicate that they are radical ions. This means
that the molecular ion contains one unpaired electron. Normally, all the electrons in a
molecule are paired, but when one electron is removed, one of the remaining electrons
is unpaired.
Generally, most molecular ions formed by electron ionization are unstable. Many of them
will immediately decompose to form fragment ions. In cases where virtually all molecular
ions decompose to form fragment ions, no molecular ions are found in the corresponding
mass spectrum. In such cases, it is difficult to determine the molecular weight of the analyte.
An example of this is shown in Figure 16.5 (upper spectrum). In the mass spectrum of
amphetamine (central stimulant), there is no peak at mass 135 corresponding to the
molecular weight.
In other cases, molecular ions are more stable and thus exist in the mass spectrum. In such
cases, the molecular weight of the compound can be read directly from the mass spectrum.
The molecular weight corresponds to one of the highest masses observed in the mass
spectrum. An example of this is shown in Figure 16.5 (lower spectrum). The mass spectrum
of lysergide (LSD, hallucinogen) contains a strong signal at mass 323 corresponding to the
molecular ion. For virtually all organic compounds, most of the molecular ions decompose
to fragment ions under electron ionization.
16.4 Identification using Electron Ionization Spectra
Mass spectra of a given substance acquired in different laboratories with different instruments are normally very similar when electron ionization is used at 70 eV. Because of this,
mass spectra are ideal for comparison, and an unknown compound can often be identified by
comparing its mass spectrum with a reference spectrum for a known compound. If the
spectrum of the unknown compound is identical to the reference spectrum, the substance
can be identified with a relatively high degree of certainty.
238
Introduction to Pharmaceutical Chemical Analysis
Figure 16.5 Mass spectra (electron ionization) of amphetamine and lysergide (LSD)
Today, mass spectra at 70 eV have been recorded and published for a very large number of
different drug substances, drug metabolites and other chemical compounds. These spectra
are stored in computer-based libraries. Mass spectrometers used for GC-MS today are
delivered with such libraries connected online to the instrument. After recording a mass
spectrum of a given unknown compound, you can electronically search for reference spectra
in the database. This makes it possible to make a quick and easy identification of the analyte.
Libraries containing mass spectra for more than 100 000 different organic compounds are
commercially available, and these are a very important tool for identification. In most cases,
pharmaceutical analysis is performed for analytes which previously were analyzed by MS,
and reference spectra are available and ready for use. An example of the use of a reference
spectrum for the identification of an unknown substance is shown in Box 16.1.
Mass Spectrometry
239
Box 16.1 Identification of fencamfamin in urine sample (upper
spectrum) using the reference spectrum (lower spectrum) recorded with
electron ionization at 70 eV
The electronic search for reference spectra often gives a list of several suggestions that can
match the spectrum of the unknown compound. The list is ranked according to how good the
fit is between the unknown and the reference. The comparison is based on the 5–10 most
intense ions. The proposals should be controlled by the operator to verify that all major
fragments in the reference spectrum are also present in the spectrum of the unknown
substance. In this case, the unknown can be positively identified from the reference
spectrum, and the identification can be further supported by retention times.
16.5 Characterization of Totally Unknowns using Electron
Ionization Spectra
Frequently, unknown compounds in GC-MS cannot be identified completely by mass
spectrometry because no matching spectra are found in the libraries. In such cases you have
to manually interpret the mass spectrum to get structural information (characterization).
This information is often combined with complementary data from IR and NMR spectroscopy. Interpretation of mass spectra is relatively difficult and requires detailed knowledge of
fragmentation patterns of organic compounds. This is beyond the scope of this textbook, but
some of the principles of interpretation are discussed in relation to a few examples, to give a
flavor of the field.
For the manual interpretation of mass spectra, it is common to start from the upper part of
the mass scale because the fragments with the highest masses are those most specific for the
240
Introduction to Pharmaceutical Chemical Analysis
Figure 16.6 Mass spectrum (electron ionization) of morphine (analgesic drug). The signal at
mass 286 corresponds to molecular ions with one 13 C atom (M þ1)þ, while 285 corresponds to
the actual molecular weight of the compound (M)þ
structure. First we try to locate the molecular ion (M)þ, to possibly get the molecular weight
of the unknown. With electron ionization, there are no signals at higher masses than the
molecular ion. We therefore look for the molecular ion at the right hand side of the mass
spectrum. The highest mass usually does not directly equal the molecular weight, but is
shifted one or two mass units [(M þ 1)þ or (M þ 2)þ] upwards from the molecular ion
because of 13 C isotopes and other heavier isotopes.
Figure 16.6 shows an example of how to determine the molecular weight. The highest
mass with a signal is at 286. This signal corresponds to molecular ions containing one 13 C
atom (M þ 1)þ, while the signal at mass 285 corresponds to the actual molecular weight of
the compound (M)þ. There is also a peak at mass 284, but this arises from loss of a H-radical
from the molecular ion (M 1)þ.
Ions with lower masses than the molecular ion have been formed by fragmentations from
the molecular ion. These fragmentations can provide significant structural information.
Table 16.2 gives different possible fragmentations directly from molecular ions, the
corresponding mass differences and possible structural information that can be derived
from the actual mass shifts. Please note that Table 16.2 is not complete. For real
interpretation, a more complete list should be used. This can be found in a number of
textbooks on mass spectrometry.
Use of the information that the different isotopes provide is very important in mass
spectral interpretation. In cases where you have very clean mass spectra without interference
from other substances, the 13 C isotope can be used to estimate the number of carbon atoms
in the molecular ion as well as in fragment ions. If the intensity of an ion with mass m is Im
and the intensity of the corresponding ion with one 13 C atom (mass m þ 1) is Imþ1, the
following equation is valid:
Imþ 1 ¼ n 0:011 Im
ð16:6Þ
where n is the total number of carbon atoms in the ion. This number can then be determined
by the following reorganization of equation (16.6):
n¼
Imþ 1
0:011 Im
ð16:7Þ
Mass Spectrometry
241
Table 16.2 Typical fragmentations from the molecular ion
Mass
Loss
M1
M2
M 15
M 16
M 16
M 17
M 18
M 19
M 20
M 26
M 27
M 28
M 28
M 29
M 29
M 30
M 30
M 31
M 32
M 33
M 34
M 35/37
M 36/38
M 41
M 42
M 42
M 43
M 43
M 45
M 45
M 46
M 46
M 60
M 127
H
H2
CH3
O
NH2
OH
H2O
F
HF
C2H2
C2H3
C2H4
CO
C2H5
HCO
CH2O
NO
OCH3
CH3OH
SH
H2S
Cl
HCl
C3H5
CH2CO
C3H6
C3H7
CH3CO
OC2H5
COOH
NO2
C2H5OH
CH3CO2H
I
Interpretation
a-Cleavage
Aromatic nitro compound, N-oxide
Primary amide
Carboxylic acid, tertiary alcohol
Primary alcohol
Fluoride
Fluoride
Unsubstituted aromatic hydrocarbon
Ethyl ester
Aromatic ethyl ether, n-propylketone
Ethyl ketone, a-cleavage
Aliphatic aldehyde
Aromatic methyl ether
Aromatic nitro compound
Methyl ester, aromatic methyl ester
Methyl ether
Thiol
Thiol
Chlorinated compound
Chlorinated aliphatic compound
Propyl ester
Aromatic acetate, ArNHCOCH3
Aromatic propyl ether
Propyl ketone, a-cleavage
Methyl ketone
Ethyl ester
Carboxylic acid
Aromatic nitro compound
Ethyl ether
Acetate
Iodated compound
An example of how to calculate the number of carbon atoms is shown in Box 16.2.
In addition to the 13 C isotope, isotope patterns of chlorine and bromine are also important
when interpreting mass spectra in pharmaceutical analysis. As shown in Table 16.1, chlorine
35
Cl and 37 Cl are in the ratio 3 : 1 and bromine as 79 Br and 81 Br are in the ratio 1 : 1.
The isotope patterns of these are therefore very characteristic as illustrated in Figure 16.7,
and all molecular ions and fragment ions containing Cl or Br are relatively easy to locate
(e.g., see the mass spectrum for chlorambucil in Figure 16.4).
To give a flavor of manual interpretation, a few mass spectra (electron ionization) of
pharmaceutical compounds is discussed in the following. Attention only focuses on direct
decompositions from the molecular ion, while further fragmentation products are not
discussed in this textbook.
Figure 16.8 shows the mass spectrum for p-amino-benzoic acid. For this compound we
find a strong signal for the molecular ion at mass 137. This information tells us that the
242
Introduction to Pharmaceutical Chemical Analysis
Box 16.2 Calculation of the number of carbon atoms based on the ratio
between 12 C and 13 C
From the mass spectrum of chlorambucil we read the following intensities for
the molecular ion with only 12 C isotopes (Im) and the molecular ion with one 13 C
isotope (Imþ1):
Im ¼ 10:7%
Imþ 1 ¼ 1:6%
This intensity ratio corresponds to the following number of carbon atoms:
n¼
1:6%
¼ 13:6 14
0:011 10:7%
The calculation demonstrates that, within the uncertainty of the measurement, the
molecular ion contains 14 carbon atoms.
Cl
2Cl
Br
2Br
Figure 16.7 Isotope patterns of chlorine (one and two atoms) and bromine (one and two atoms).
Two mass units between each peak
Figure 16.8 Mass spectrum (electron ionization) of p-amino-benzoic acid
Mass Spectrometry
243
Figure 16.9 Mass spectrum (electron ionization) of bromazepam
molecular ion is relatively stable, and the m/z value of the molecular ion gives us the
molecular weight of the compound. Note that the molecular ion has an odd nominal mass.
This follows the nitrogen rule, which states that if a compound has an odd number
of nitrogen atoms then the molecular ion has an odd nominal mass. If a compound has
an even number of nitrogen atoms, or no nitrogen atoms, the molecular ion has an even
nominal mass.
The next fragment of interest is found at mass 120, which makes a mass difference of 17
compared to the molecular ion. According to Table 16.2 this is corresponding to the loss of
an OH radical (M 17)þ, which is frequently observed for carboxylic acids and tertiary
alcohols. We also find a (M 45)þ fragment at mass 92 corresponding to the loss of a
COOH-radical, and this confirms that the compound is a carboxylic acid.
A number of drugs contain halogen atoms (F, Cl, Br, I). According to Table 16.2, such
compounds may give characteristic release of the halogens corresponding to (M 19)þ for
F, (M 35)þ for Cl, (M 79)þ for Br and (M 127)þ for I. An example of the release of
bromine is shown in Figure 16.9 for bromazepam (tranquilizer).
The molecular ion for bromazepam is found at mass 315. This is consistent with the
nitrogen rule, since bromazepam contains three nitrogen atoms. The compound contains a
bromine atom, and the typical isotope distribution of bromine can be seen clearly (compare
with Figure 16.7); the heavier bromine isotope 81 Br gives a peak at mass 317 with about the
same intensity as the peak at 315. Normally, bromine is easily released from the molecular
ion as a CBr bond in general is relatively weak compared to CF and CCl bonds. For
bromazepam we can see a peak at (M 79)þ, corresponding to mass 236. The fragment ion
at mass 236 no longer contains bromine, and no bromine isotope pattern is observed at this
fragment. Mass 236 is a large peak in the spectrum which says that the release of bromine
from the molecular ion in this case is a dominant process. The isotope pattern at the
molecular ion and the (M 79)þ decomposition tells us clearly that this compound contains
bromine.
An example of the release of chlorine is shown in Figure 16.10 for clonazepam
(anticonvulsant). Clonazepam contains three nitrogen atoms, and the molecular ion is
located at mass 315. Clonazepam contains one chlorine atom, and a signal with one-third
intensity of 315 is observed at 317 corresponding to the 37 Cl isotope. This gives a strong
indication of the presence of chlorine. In addition, a strong signal is observed at m/z 280, and
this signal is due to the loss of chlorine from the molecular ion. Observe that the chlorine
244
Introduction to Pharmaceutical Chemical Analysis
Figure 16.10 Mass spectrum (electron ionization) of clonazepam
isotope pattern is not present at m/z 280. This tells us that a single chlorine atom is present in
the structure.
A strong signal is also observed at m/z 314, and this corresponds to the loss of a hydrogen
radical from the molecular ion. We also see a strong signal at mass 286, and this corresponds
to the loss of CHO from the molecular ion.
16.6 Chemical Ionization
As we have seen in Section 16.5, the molecular ion may be unstable during electron
ionization in GC-MS and may totally decompose to fragment ions. In such cases, the
molecular ion is not present in the mass spectrum, and it is not possible to establish the
molecular weight of the compound. For identification, the lack of molecular ions is a
significant disadvantage. If electron ionization does not provide sufficient molecular ions, it
means that the electrons that bombard the sample contain too much energy. The high energy
is causing all the molecular ions to form fragment ions. This can be circumvented by
reducing the energy of the electrons, and this is done by adding a reagent gas into the ion
source. This is known as chemical ionization (CI). Chemical ionization, which takes place in
the vacuum region of the mass spectrometer, is normally performed when the MS is working
as a detector for GC (GC-MS). Methane is a typical reagent gas used for chemical
ionization. The reagent gas is supplied in relatively large amounts compared to the amount
of sample (1000–10 000 times more reagent gas than sample). Electrons released by the
rhenium or tungsten filament in the ion source therefore almost exclusively ionize the
reagent gas. With methane as reagent gas, CH5þ and C2H5þ are formed. These ions are
highly reactive and react with the analyte (XH) according to the following reactions:
þ
CHþ
5 þ XH ! XH2 þ CH4
ð16:8Þ
þ
C2 Hþ
5 þ XH ! XH2 þ C2 H4
ð16:9Þ
þ
C2 Hþ
5 þ XH ! X þ C2 H6
ð16:10Þ
Mass Spectrometry
245
44
Electron ionization
50
150
100
m/z
136
Chemical ionization
50
100
150
m/z
Figure 16.11 Mass spectra (electron ionization and chemical ionization) of amphetamine
In Equations (16.8) and (16.9), the chemical ionization gives a molecular ion with mass
(M þ 1)þ. In the lower reaction [Equation (16.10)], the chemical ionization gives a
molecular ion with mass (M 1)þ. As you can see, chemical ionization is accomplished
with ions, while electron ionization is accomplished with electrons.
Chemical ionization with methane normally form either ions with mass (M þ 1)þ or
(M 1)þ. These ions are relatively stable, and therefore they appear as intense signals in the
mass spectra. Because of this, chemical ionization is very powerful for the determination of
molecular masses. An example of this is illustrated in Figure 16.11, with the mass spectra for
amphetamine both with electron ionization and chemical ionization.
In the electron ionization spectrum of amphetamine, it is obvious that all the molecular
ions have fragmented, mainly to C2H6Nþ with mass 44, and there is no signal detected at
mass 135 corresponding to molecular weight. Using chemical ionization with methane,
however, you get the base peak at mass 136. This corresponds to (M þ 1)þ, and the
molecular weight of the compound may be determined to 135.
CI is usually carried out with methane as reagent gas. In addition to methane, we can also
use other gases, such as iso-butane and ammonia. Iso-butane and ammonia are used in cases
246
Introduction to Pharmaceutical Chemical Analysis
where less fragmentation is requested. These reagents bind Hþ more strongly than CH4 does
and impart less energy to MHþ when the proton is transferred to M. In addition to this you
can also in CI take advantage of that both positive and negative ions can be produced.
Compounds containing electronegative atoms form significant amounts of negative ions,
and these ions may be analyzed with high sensitivity and specificity by negative chemical
ionization mass spectrometry.
16.7 Electrospray Ionization
Another extremely important ionization technique in mass spectrometry is electrospray
ionization (ESI). Electrospray ionization is very much in use as an ionization technique
when MS is used for detection in HPLC. HPLC coupled with MS is normally abbreviated
LC-MS. With ESI, the sample is delivered to the mass spectrometer in a liquid stream with
relatively high volatility, which is normally the mobile phase from an HPLC system. The
principle for the ionization, which in this case takes place in atmospheric pressure outside
the vacuum region of the mass spectrometer, is shown in Figure 16.12.
The mobile phase from the HPLC column passes through a narrow capillary. At the end of
this capillary, a fine aerosol is formed by nitrogen gas flowing along the tip of the capillary.
The aerosol consists of a large number of very small droplets of mobile phase containing the
analyte. A cylindrical electrode is placed in continuation of the capillary. An electrical
potential of typically 2–5 kV is coupled between the capillary tip and the cylindrical
electrode. This potential gives electrical potential to the small drops. Because the mobile
phase is volatile, the liquid droplets evaporate and are flushed away by a drying gas. The
analyte molecules remain charged and are extracted into the vacuum area of the mass
spectrometer for further analysis by the mass analyzer. Three essential differences between
electron ionization (EI) and electrospray ionization (ESI) should be pointed out:
.
ESI takes place under atmospheric pressure outside the mass spectrometer. EI takes place
under vacuum inside the mass spectrometer.
Cylindrical electrode
High voltage
2-5 kV
Drying gas
N2
Mobile phase
from HPLC
Mass analyzer
(vacuum)
Narrow capillary
Figure 16.12 Principle of electrospray ionization
Mass Spectrometry
247
ESI is a soft ionization technique producing mainly (M þ 1)þ or (M 1)þ. EI is a hard
ionization technique producing substantial amounts of fragment ions in addition to
molecular ions Mþ.
. ESI is used for LC-MS. EI is used for GC-MS.
.
Electrospray ionization mainly gives (M þ 1)þ or (M 1)þ ions, which are relatively stable
and which do not decompose to fragment ions. Thus, mass spectra following EI essentially
contain (M þ 1)þ or (M 1)þ. Electrospray ionization is used for acidic or basic analytes;
and to get high signals in the mass spectrometer, it is important that the analytes are ionized
in the mobile phase from the HPLC system. This means that the pH of the mobile phase must
be adjusted so that the analytes are charged. For basic drugs, this is accomplished having
acidic conditions in the mobile phase. In addition, it is important that the mobile phase is
made with volatile components, so that they are transported away as efficiently as possible
with the drying gas. This means that volatile buffer components should be used in the mobile
phase, such as ammonium formate, formic acid, or ammonia, and that nonvolatile buffer
components like phosphates should be avoided.
16.8 Atmospheric Pressure Chemical Ionization
In addition to electrospray ionization, it is common to use atmospheric pressure chemical
ionization (APCI) when MS is coupled with HPLC (LC-MS). Also in APCI the sample is
delivered to the mass spectrometer in a liquid flow (mobile phase from the HPLC), and the
ionization is performed outside the mass spectrometer at atmospheric pressure. Ions pass
further into the vacuum in the mass spectrometer for the final mass determination. The
principle of APCI is shown in Figure 16.13.
The mobile phase containing the analytes flows through a heated capillary, and at the end
of the capillary the mobile phase is evaporated and transported away by a drying gas. A
needle connected to a high electric potential of the order of several kilovolts ensures that
Drying gas
Needle with high
electrical potential
N2
Mobile phase
from HPLC
H3O+
Mass analyzer
(vacuum)
Heated capillary
Figure 16.13 Principle of atmospheric pressure chemical ionization
Introduction to Pharmaceutical Chemical Analysis
248
nitrogen and small amounts of water that exist at the end of the capillary are ionized
according to the following reactions:
N2 þ e ! N2þ þ 2e
ð16:11Þ
N2þ þ 2N2 ! N4þ þ N2
ð16:12Þ
N4þ þ H2 O ! H2 O þ þ 2N2
ð16:13Þ
.
.
.
.
.
H2 O þ þ H2 O ! H3 Oþ þ OH
.
þ
.
ð16:14Þ
.
H3 O and OH (OH radical) can then react with the analyte (MH) according to the following
reactions:
H3 Oþ þ MH ! MHþ
2 þ H2 O
ð16:15Þ
OH þ MH ! M þ H2 O
ð16:16Þ
.
Protonated molecular ions (MHþ
2 ) or deprotonated molecular ions (M ) are introduced into
the mass spectrometer, and the masses are determined. Also, APCI is a relatively soft
ionization technique, and mainly molecular ions are formed. Atmospheric pressure
chemical ionization can be used as an alternative to electrospray ionization and is
particularly relevant if the analytes do not contain acidic or basic groups. For this type
of compounds, it can be difficult to get sufficient signals with EI.
16.9 High-Resolution Mass Spectrometry
Until now we have principally discussed mass spectrometry with low resolution, which
means that all the masses have been rounded to the nearest integer. In some cases, the use of
nominal masses is impractical because important structural information can be lost. An
example of this is shown in Box 16.3: a molecular ion with nominal mass 120 can
correspond to several different types of ions that mutually have very small differences
in the exact mass. The elemental compositions of the various ions at mass 120 are very
Box 16.3 Several ions with the same nominal mass but with significant
differences in their exact mass
Ion
C5H4N4þ
C7H8N2þ
C9H12þ
C8H8Oþ
Nominal mass
Exact mass
120
120
120
120
120.044
120.069
120.096
120.058
Mass Spectrometry
249
different, and for correct interpretation of the mass spectrum it may be essential to know the
correct elemental composition. Elemental compositions can be determined if the mass
determinations are highly accurate. This is done by MS with high resolution, where the mass
spectrometer is tuned for highly accurate mass determinations. If the mass of the molecular
ion in Box 16.3 is determined to 120.070 ( 0.0005), it can be concluded that this
corresponds to the elemental composition of the ion C7H8N2þ.
High resolution MS can thus determine the elemental composition of both molecular ions
and fragment ions.
High resolution MS requires highly sophisticated instrumentation, and not all mass
spectrometers can be used for accurate mass determinations. Thus, different mass spectrometers have different mass resolutions. Mass resolution can be quantified by the resolving
power, characterized by the ability to separate two ions of equal intensity with m/z values
differing by Dm and with no more that 10% overlap in their m/z signals:
m
Resolving power ¼
ð16:17Þ
Dm
where m is the smaller value of m/z. The resolving power is an important value to
characterize the performance of a mass spectrometer. An example on how to calculate
resolving power is shown in Box 16.4.
Box 16.4 Calculation of resolving power
A mass spectrometer is capable of separating m/z 120.044 and m/z 120.058 from
Box 16.3 with approximately 10% overlap as illustrated in the figure below.
120.058
120.044
m/z
This corresponds to the following resolving power:
Resolving power ¼
m
120:044
¼
¼ 8575 8600
Dm 120:058 120:044
The resolving power of the mass spectrometer is thus about 8600.
250
Introduction to Pharmaceutical Chemical Analysis
16.10 Instrumentation
The instrument used for MS is called a mass spectrometer. This is a very expensive
instrument. A general overview of a mass spectrometer is illustrated in Figure 16.14.
The first step in MS is the ionization. As previously mentioned, ionization can be based on
electron ionization (EI) or chemical ionization (CI), and in both cases the ionization occurs
under vacuum in an ion source inside the mass spectrometer. This is typical for GC-MS.
However, for electrospray ionization (ESI) and atmospheric pressure chemical ionization
(APCI) the ionization takes place outside the vacuum region of the instrument. This is
typical for LC-MS. In all cases, the molecular ions and the fragment ions fly into the
mass analyzer where they are separated according to their mass to charge ratio (m/z). After
the mass analyzer, the different ions are detected with the detector. The results are processed
electronically by a computer which provides the basis for the recording and plotting of
mass spectra.
Mass spectrometry is performed under vacuum conditions, with a very low pressure
inside the instrument. The reason for this is to prevent the ions of interest from colliding with
air molecules. This ensures that the ions reach the detector. To maintain a low pressure in the
mass spectrometer, the instrument must be closed, and air penetrating the system must
constantly be pumped out of the system by means of powerful pumps. The pumps are
connected to both the ion source (in the case of EI and CI) to the mass analyzer and to the
detector, and this ensures that the pressure inside the instrument does not exceed 104 to
108 torr (107 to 1011 bar).
As discussed above, it is usual to use the mass spectrometer as a chromatographic detector
in pharmaceutical analysis (GC-MS and LC-MS). GC-MS is done today mainly with
capillary columns, and the outlet of the capillary column is placed directly inside the ion
source of the mass spectrometer. Thus, electron ionization or chemical ionization takes
place directly when the compounds leave the GC column. The ion source is a small chamber
inside the MS system just before the mass analyzer. Usually 1–3 ml/min of carrier gas
(helium or hydrogen) is used in capillary GC, and this gas also enters the mass spectrometer.
However, the capacity of the pumps that evacuate the instrument is so powerful that
the pressure does not rise significantly due to the carrier gas coming from the GC column. In
LC-MS the source for electrospray ionization or atmospheric pressure chemical ionization
is situated outside the mass spectrometer. After ionization ions pass through a very narrow
Vacuum with EI and CI
Atmospheric pressure with ESI and API
Sample
introduction
Ion source
Mass analyzer
Detector
Computer
Pump
Figure 16.14 Schematic overview of a mass spectrometer
Mass Spectrometry
251
Detector
Ion source
RF and dc
voltages
Figure 16.15 Quadrupole mass spectrometry
opening into the mass spectrometer and further into the mass analyzer. LC-MS is somewhat
complicated by the fact that the mobile phase has to be evaporated and removed before the
analyte enters the vacuum zone of the mass spectrometer. If this is not performed, a mobile
phase flow at 1 ml/min produces about 1 l/min of gas by evaporation inside the mass
spectrometer, which causes the pressure to rise dramatically. This problem is partly solved
by using the drying gas to remove much of the mobile phase (as discussed above), partly by
using narrow-bore HPLC columns and partly by splitting the mobile phase flow from the
column. By splitting the mobile phase flow, only a small portion of the fluid passes into the
mass spectrometer. In addition, powerful pumps are used in LC-MS to keep the vacuum
inside the instrument.
In the mass analyzer the ions are separated according to their m/z values. Many
instruments in use today use a quadrupole mass analyzer. The quadrupole mass analyzer
is based on four parallel rods to which are applied both a constant voltage and a radio
frequency oscillating voltage (Figure 16.15). The electric field deflects ions in complex
trajectories as they pass through the mass analyzer, allowing only ions with one particular
m/z ratio to reach the detector. Other ions collide with the rods and are lost before they reach
the detector. Rapidly varying voltages select ions of different m/z values to reach the
detector, and in this way the instrument can scan a wide m/z range in a very short time.
Actually, quadrupole instruments can record 2–8 spectra/s in the mass range up to m/z 4000.
Quadrupole instruments are popular as they are reasonable in price, but they are not
sufficiently accurate to enable for high resolution mass spectrometry.
In addition to quadrupole mass analyzers, also ion-trap analyzers and time of flight
analyzers are used frequently in pharmaceutical analysis. In ion-trap MS the substances to
be analyzed are introduced directly into the mass analyzer, usually from a chromatographic
column. The mass analyzer is called an ion trap and is in principle a small circular chamber
as illustrated in Figure 16.16.
Ions generated either by electron ionization or electrospray ionization are collected in the
ion trap. The ion trap is surrounded by a ring electrode, to which a constant-frequency radio
frequency voltage is applied. This causes the ionized substances to move in stable
trajectories inside the ion trap. Increasing the amplitude of the radio frequency voltage
expels ions of a particular m/z value by sending them into unstable trajectories that pass
through the exit holes in the end caps. Ions expelled through the lower end cap are captured
by the detector and registered with high sensitivity. Ion-trap MS is a compact instrument that
is very well suited for chromatographic detection.
252
Introduction to Pharmaceutical Chemical Analysis
Filament
End cap
eeRing electrode
+
+
Ring electrode
+
Chromatography
column
End cap
Detector
Figure 16.16 Ion-trap mass spectrometry
In time of flight MS, ions formed in the ion source are accelerated in pulses by means of an
electric potential imposed on a back plate right in the back of the ion source. Positive ions are
accelerated by a positive potential applied to the back plate. Ions fly from the ion source and
into a field-free flight tube without any magnetic or electrostatic field as illustrated in
Figure 16.17.
All ions are accelerated to the same kinetic energyð1=2 mv2 Þ, which means that the
heavier ions will move more slowly than light ions. At the other end of the tube is a detector.
The light ions will hit the detector at the other end of the flight tube faster than heavier ions
(lower flight time), which can be used to measure the mass of the various ions. The mass of
each ion is thus determined based on its flight time. The advantages of time of flight MS is
that it can be used to measure ions in a very large mass range, the spectra can be taken up very
quickly and the instruments can under certain conditions be used for mass spectrometry with
high resolution.
Another possibility to make high resolution mass spectrometry is to use a double focusing
mass spectrometer. The principle of this type of instrument is shown in Figure 16.18.
First the ions pass through an electrostatic analyzer. The electrostatic analyzer consists of
two metal plates applied to an electrostatic field. The inner plate is negatively charged, while
the outer plate is given a positive potential. This means that positive ions from the ion source
are attracted towards the negative plate, and the deflection will depend on the strength of the
electrostatic field applied to the plates and on the m/z value of the ions. The electrostatic
analyzer focuses ions with the same mass, even if these have a slightly different speed and
Ion source
+
+
+
Flight tube
+
+
+
+
Detector
+
+
Back-plate
Figure 16.17 Time of flight mass spectrometry
Mass Spectrometry
253
+
Electrostatic
analyzer +
-
Magnetic analyzer
-
+
-
Ion source
Detector
Figure 16.18 Double focusing mass spectrometry
direction from ion source. The electrostatic field can be varied in order to scan ions with
different mass through the electrostatic analyzer. Then the ions enter a magnetic analyzer in
which the ions enter a magnetic field which causes deflection of their trajectories depending
on the magnetic field strength and the m/z values of the ions. By varying the magnetic field,
different ions pass the magnetic analyzer and are registered by the detector. The focusing in
the electrostatic analyzer makes the mass determination in the subsequent magnetic
analyzer much more accurate, so that the instrument can distinguish between different
ions with very small difference in mass (exact mass).
Mass spectra can be acquired very rapidly, typically 1–20 times every second. Since MS is
usually combined with chromatography, it is clear that a lot of data are generated. Because of
the large volume of data generated, it is necessary to connect mass spectrometers online to
computers with a good storage capacity for the data. This provides great flexibility, and mass
spectra from GC-MS and LC-MS analysis can be printed for analysis and examined with
regard to identification and quantification. In computers for GC-MS, it is normal to instal a
library of reference spectra, and the spectra of unknown compounds can be automatically
identified by an electronic search for reference spectra.
16.11 Chromatography Coupled with Mass Spectrometry
Mass spectrometry is in most cases carried out as GC-MS or LC-MS. In both cases sample
constituents are separated by passage through a chromatographic column. After the
chromatographic column, each compound elutes into the mass spectrometer for mass
spectrometric analysis. The mass spectrometer is used in this manner as an advanced
chromatography detector. Mass spectrometers in combination with chromatography may be
operated according to several different principles, of which three are discussed in this text:
.
Full scan and recording of spectra;
Selected ion monitoring (SIM);
. Selected reaction monitoring (SRM).
.
254
Introduction to Pharmaceutical Chemical Analysis
Total ion-current
Total ion-current chromatogram
2
4
1
3
Retention time
Intensity
Mass spectrum of compound 3
m/z
Figure 16.19 Total ion current chromatogram and mass spectrum for component 3 in a mixture
of four components
If you want as much structural information about the analytes as possible, it is common to
operate the mass spectrometer in the full scan mode where the instrument continuously
records full mass spectra within a given mass range. Because the separated substances are in
the mass spectrometer only for a very short time, it is important that mass spectra are
recorded with very short time intervals. Thus, typically 1–5 mass spectra are recorded per
second. The mass spectra are stored continuously during the analysis in a computer, and the
results can be plotted as a chromatogram (detector signal vs retention time) showing a peak
for each of the separated and ionized substances. The separated substances can be identified
or characterized by examining their mass spectra stored on the computer. The total amount
of ions (total ion current; TIC), or the sum of all ions, in each mass spectrum is plotted
continuously as function of separation time, resulting in a total ion current chromatogram
(see Figure 16.19).
The total ion current is very low when only the carrier gas/mobile phase flows into the
mass spectrometer. Each time a compound elutes into the mass spectrometer, the total ion
current increases for a short period of time, giving rise to a peak in the total ion current
chromatogram. By recording the mass spectra, a lot of structural information is available
for each of the separated substances, but the drawback is that sensitivity is limited in
the full scan mode, and it may therefore be difficult to detect compounds at very low
concentration levels.
If you want to analyze substances at very low concentration levels or if you want to
perform quantitative measurements, it is common to operate the mass spectrometer in
Mass Spectrometry
255
Ion signal m/z 286
Selected ion chromatogram
m/z 286
N-Desmethylclobazam
9.5 min
Retention time
Figure 16.20 Selected ion chromatogram (mass 286) for GC-MS analysis of human plasma.
N-Desmethylclobazam eluted into the mass spectrometer at 9.5 min (¼ retention time) and
formed ions with mass 286, resulting in the peak observed in the selected ion chromatogram.
Plasma samples contains a large number of other substances but these do not appear because
their ionization/fragmentation does not form ions with mass 286
selected ion monitoring mode (SIM). In SIM, the mass spectrometer is tuned to measure
only a single mass or a few selected masses during the chromatographic separation. From
SIM analysis you get a selected ion chromatogram, which is a plot of the intensity of the
particular mass as a function of retention time. Components in the sample that give the
requested mass during ionization and fragmentation would therefore be found as peaks in
the selected ion chromatogram (see Figure 16.20), while other compounds that do not form
ions with that particular mass are not displayed.
If SIM does not give adequate sensitivity or specificity, one can take advantage of selected
reaction monitoring (SRM). SRM analysis can be accomplished with a triple quadrupole
mass spectrometer, where the mass analyzer consists of three quadrupoles in series (see
Figure 16.21).
The ions from the ion source enter the first quadrupole (Q1). This quadrupole is locked to
a mass that is characteristic of the substance to be determined, which typically can be the
molecular ion. This ion which is selected in the first quadrupole is called the precursor ion.
All other masses can not pass the first quadrupole and this gives rise to the high selectivity of
SRM. The precursor ions then pass to the next quadrupole (Q2), where they react with either
N2 or Ar. The configuration of the second quadrupole can vary from instrument to
instrument, and in some cases it can be either a hexapole (six rods) or an octapole
Formation of product ions
Measurement of product ions
Selection of precursor ion
Detector
Ion source
Q1
Q2
Q3
N2 or Ar
Figure 16.21 Triple quadrupole mass spectrometer
256
Introduction to Pharmaceutical Chemical Analysis
Ion signal m/z 182.2
Selected reaction monitoring chromatogram
m/z 304.1
m/z 182.2
Cocaine
3.6 min
Retention time
Figure 16.22 Selected reaction monitoring chromatogram for LC-MS analysis of cocaine.
Cocaine eluted into the mass spectrometer after 3.6 min. Mass 304.1 was selected as precursor
ion and mass 182.2 was measured as product ion
(eight rods). Here, the ions fragment to product ions, and the product ions are measured with
the third quadrupole (Q3) and with the detector. Product ions are usually very specific to a
given substance, and detection of product ions at given masses therefore provides a highly
reliable identification. Such measurements can be made with very high sensitivity because
the ion background from other substances is effectively eliminated. Figure 16.22 demonstrates a chromatogram from a SRM analysis.
Triple quadrupole mass spectrometers are increasingly being implemented in the
pharmaceutical laboratories due to their high sensitivity and specificity.
16.12 Quantitative GC-MS and LC-MS
As we discussed earlier, mass spectrometry is very suitable for identification. Like other
spectroscopic methods used in pharmaceutical analysis, mass spectrometry can also be used
for quantitative analysis. Quantitative analysis is based on the fact that areas under an
Peak area @ m/z 286
Peak area in unknown sample
= Point for standard
solution
Concentration in unknown sample
Plasma concentration of N-desmethylclobazam
Figure 16.23 Calibration curve for quantitative determination of N-desmethylclobazam in
plasma, based on six standard solutions with different concentrations of N-desmethylclobazam
Mass Spectrometry
257
analyte peak in a selected ion chromatogram (or a selected reaction monitoring chromatogram) are proportional to the concentration of the analyte. By calibration with standard
solutions of known concentration, one can plot the standard curves of peak area as a function
of concentration, and from these one can determine the amount of the substance in unknown
samples. This is similar to the quantification by other chromatography detectors.
To quantify N-desmethylclobazam in plasma using SIM at mass 286 (see Figure 16.20)
a calibration curve has to be established. Drug-free plasma is added to known amounts of
N-desmethylclobazam and analyzed by GC-MS. The peak area in the selected ion
chromatogram for mass 286 is plotted as a function of concentration (Figure 16.23).
The standard curve can be used for GC-MS of unknown plasma samples to determine
concentrations of the drug substance. Quantification using GC-MS and LC-MS with
selected ion monitoring is widely used in pharmaceutical analysis.
16.13 Areas of Usage and Performance
Within pharmaceutical analysis, mass spectrometry is mainly used for the determination of
drug substances and their metabolites in blood and urine. Because these sample types are
very complex, the mass spectrometric analysis is normally done in combination with
chromatography so that the different components are separated from each other prior to
mass spectrometry. Gas chromatography/mass spectrometry (GC-MS) is an extremely
important technique both within therapeutic drug monitoring (TDM), illicit drug analysis
and in doping analysis. LC-MS is also a very important technique in pharmaceutical
analysis for therapeutic drug monitoring (TDM), illicit drug analysis and for drug analysis in
pharmaceutical research.
Mass spectrometry is a very sensitive analytical method and can detect compounds down
to the pico- or femtogram level. This is especially important for GC-MS and LC-MS
analysis of blood and urine samples, where drug substances and metabolites can be present
in very low concentrations and where the amount of sample can be limited. Compared with
other analytical methods, MS (and GC-MS and LC-MS) is complicated and requires
expensive instrumentation. The costs associated with mass spectrometric analysis are
therefore relatively high. Operation of mass spectrometers and interpretation of data both
requires high skills for the laboratory personal.
16.14 Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry
Matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) is a soft MS
ionization technique which is primarily used for the analysis of large biomolecules, like
proteins, peptides and sugars. Ionization is accomplished with a laser beam. Typically, a
small solution containing the analyte is mixed with a solution of a matrix component such
as 2,5-dihydroxybenzoic acid. The liquid is evaporated, providing a dry spot of fine
crystals of the matrix plus the analyte. The laser beam is directed towards this spot. The
matrix absorbs the laser energy, and primarily the matrix is ionized. The matrix then
transfers part of its energy and charge to the analyte molecules, while protecting them
258
Introduction to Pharmaceutical Chemical Analysis
from the disruptive energy of the laser. The charged analytes are finally analyzed in the
mass spectrometer, and typically time of flight analyzers are used for this purpose.
MALDI-MS is an official technique in the European Pharmacopoeia, and the use of this
technique is expected to increase in the future with the introduction of more drugs based on
biological macromolecules.
16.15 Inductively Coupled Plasma Mass Spectrometry
An inductively coupled plasma (ICP), as discussed in Chapter 9, can be used as an ion source
in mass spectrometry, and this technique is called inductively coupled plasma mass
spectrometry (ICP-MS). Samples of interest are introduced into the ICP. The ICP, as
discussed earlier, is a hot plasma (atmosphere) of argon with a temperature in the order of
10 000 K. Here the sample constituents are atomized into free atoms, and some of the free
atoms are also ionized by losing an electron. From the ICP, the ionized atoms are transferred
into a mass spectrometer where they are measured and detected based on their m/z values. A
schematic illustration of an instrument for ICP-MS is illustrated in Figure 16.24.
An important part of the ICP-MS instrument is the interface that couples the ICP, which
operates at atmospheric pressure, and the mass spectrometer which operates at vacuum
conditions. First, the hot plasma gas with the ionized atoms from the ICP are transferred
through a cone with a very small orifice (<1.0 mm). This cone is located at the entrance of
the mass spectrometer. On the other side of the cone, the pressure is at 1 torr and this pressure
is maintained by a mechanical pump. A fraction of the plasma gas is then passed through a
second cone (skimmer) and into a chamber with the same pressure as in the rest of the mass
spectrometer (<10–4 torr). Here the positive ions of interest are separated from electrons and
molecular species by a negative voltage and then accelerated and focused by a magnetic
ion lens into the quadrupole mass analyzer. The quadrupole mass analyzer serve as a mass
filter in the same way as discussed in Section 16.10, and the different atomic ions with
different m/z values are detected by a detector located after the quadrupole mass analyzer.
ICP-MS is used for the determination of (atomic) elements in the same way as AAS and
AES (discussed in Chapter 9). ICP-MS is typically operated in the mass range of m/z 3–300.
More than 90% of the elements in the periodic table have been determined by ICP-MS. The
1 torr
<10-4 torr
Ion lenses
Cone
Quadrupole
Sample
Detector
ICP
Skimmer
Pumping
Figure 16.24 Schematic illustration of an ICP-MS instrument
Mass Spectrometry
259
spectra produced by ICP-MS are remarkably simple compared to conventional ICP optical
spectra, and consist of a simple series of isotope peaks for each element present in the
plasma. These spectra are used to identify elements present in the sample and for their
quantitative determination. Usually, quantitative analyses are based on calibration curves in
which the ratio of the ion count for the analyte to the count for an internal standard is plotted
as a function of concentration. ICP-MS is a highly sensitive instrument, and elements can be
detected down to the 0.1–10.0 ppb level. The use of ICP-MS is expected to increase in the
pharmaceutical quality control laboratory in the future, among others for the trace level
detection of foreign elements in pharmaceutical raw materials (discussed in detail in
Chapter 21).
16.16 Summary
Mass spectrometry is a key method in pharmaceutical analysis and is used both in
quantitative analysis and for identification purposes. The technique is mostly used in
conjunction with analysis of drugs in biological samples. In combination with GC, it is
common to ionize the analytes under vacuum conditions, either by EI (in which the analytes
are bombarded with electrons) or by CI (where the analytes reacts with charged reagent gas
ions). The molecular ions formed then completely or partially fragment into ions with lower
masses, which are characteristic for each compound. The ions are separated according to
their m/z value in a mass analyzer and detected by a detector. This process forms the basis for
the plotting of mass spectra. In combination with LC, ionization is performed at atmospheric
pressure either by electrospray ionization or atmospheric pressure chemical ionization.
Unknown compounds are identified from their corresponding mass spectra, while they are
quantified on the basis of peak areas in selected ion chromatograms.
17
Miscellaneous Chemical Techniques
The preceding chapters discuss a broad range of chemical techniques, like high performance liquid chromatography (HPLC), titration, and IR spectroscopy. These techniques are
very important for the quality control of pharmaceuticals and for bioanalysis, and therefore
they are treated very comprehensively. In addition to these principal techniques, several
other techniques are in use for pharmaceutical analysis. This chapter discusses some of
these other techniques. Priority is given to techniques officially adopted by the European
Pharmacopoeia, and the principles are discussed only briefly. For more comprehensive
discussions, the reader has to look in other textbooks.
17.1 Potentiometric Determination of Ions using Ion-Selective Electrodes
A preceding chapter about titration (Chapter 5) discusses potentiometric endpoint detection
with pH-sensitive glass electrodes. In this set-up, pH is measured continuously in the
titration vessel during the titration process. This is accomplished by measuring the potential
difference between a reference electrode and a pH-sensitive glass electrode. The potential of
the reference electrode is constant, whereas the potential of the pH-sensitive glass electrode
is dependant on the pH in the vessel.
The pH-sensitive glass electrode is one example of an ion-selective electrode, which
responds selectively to a certain type of ion. However, other ion-selective electrodes are also
available, which respond selectively to other types of ions, like potassium ions (Kþ) and
fluoride ions (F). Ion-selective electrodes can be liquid-based ion-selective electrodes or
solid-state ion-selective electrodes. Figure 17.1 illustrates the general principle of an ionselective electrode.
The ion-selective electrode in Figure 17.1 consists of an internal reference electrode, an
internal filling solution and an ion-selective membrane. The ion-selective membrane is
capable of binding only the intended type of ions. The ion-selective membrane can be a
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
262
Introduction to Pharmaceutical Chemical Analysis
To potensiometer
Internal reference
electrode
External reference
electrode
Internal filling
solution
Ion-selective
membrane
Sample solution
Figure 17.1 General principle of an ion-selective electrode
polymeric film providing a liquid-based ion-selective electrode, or it can be an inorganic
salt crystal providing a solid-state ion-selective electrode. Symbolizing the intended type
of ions as Cþ, the membrane contains a certain type of molecules that selectively binds to
Cþ. These molecules are called the ionophore. The ionophore is selected to have a high
affinity towards Cþ, and ideally with no affinity for other ions. Real electrodes however
always have some affinity for other ions, and these can interfere with the measurement.
The internal filling solution contains Cþ ions, and a certain and constant amount of
these are attached to the membrane through binding to the ionophore. Therefore, the
surface charge on the inner surface of the membrane is constant. On the outer surface of the
membrane, which is in contact with the sample solution, the amount of Cþ ions attached to
the membrane is dependant on the concentration of Cþ in the sample. Thus, a different
surface charge develops on the outside of the membrane, and this difference in surface
charge over the membrane can be measured by the internal reference electrode. The
potential of the ion-selective electrode always has to be measured relative to an external
reference electrode having a constant potential. Typically, either a silver–silver chloride
electrode or a calomel electrode is used as reference electrode. These reference electrodes
are described in Chapter 5.
Ion-selective electrodes respond linearly to the logarithm of analyte concentration (or
more correctly analyte activity) over four to six orders of magnitude. By using a calibration
curve, the measured potential difference can be related to the analyte concentration.
However, it is important that the composition of the standard solution used to make the
calibration curve closely approximates the composition of the unknown samples, because
the sample matrix can affect the measurement. Therefore, standard addition is frequently
used in combination with ion-selective electrodes, to avoid interferences from the matrix.
As discussed earlier, the sample is measured without and with standard solution added to the
sample during standard addition. The increase in response after addition of an exactly
known amount of standard to the sample is used for establishing the calibration curve.
Potentiometric determination of ionic concentrations using ion-selective electrodes is
used in some cases for the quality control of active pharmaceutical ingredients, excipients
and final pharmaceutical products. One example is for the assay of sodium fluoride in the
Miscellaneous Chemical Techniques
263
European Pharmacopoeia, where the use of a fluoride-selective electrode is prescribed. Ionselective electrodes can also be used for measurements in biological fluids and are
frequently used to measure the important blood parameters Naþ, Kþ, Cl and total CO2
in hospital laboratories.
17.2 Paper Chromatography
In a preceding chapter (Chapter 12), thin-layer chromatography (TLC) is discussed in detail.
Paper chromatography is essentially very similar to TLC. In paper chromatography,
the stationary phase is a piece of absorbent paper. The sample is spotted on a line
drawn on the paper (starting line), and standards can also be spotted on the same line as
illustrated in Figure 17.2.
The paper is then suspended in a container with a layer of mobile phase in the bottom. The
container has to be closed to make sure that the atmosphere in the container is saturated with
the mobile phase. Saturation of the atmosphere in the container with vapor stops solvent
from evaporating as it rises up the paper. As the mobile phase slowly travels up the paper by
capillary forces, the different constituents of the sample travel with different rates and the
sample is gradually separated into different spots, representing the individual constituents
of the sample. After a certain period of time, the paper chromatography is finished by
removing the paper from the container. The position of the mobile phase front has to be
marked before the mobile phase evaporates from the paper. Colored components are easily
observed on the paper, whereas colorless spots have to be visualized by some type of
chemical reaction to give colored spots. This is very similar to what is discussed under TLC.
Although paper chromatography is easy to perform, the retention mechanism can be
difficult to explain. Paper is made of cellulose fibers, and cellulose is a polymer of glucose.
Cellulose contains a large number of OH groups, and these groups play an important role
during separation. Cellulose fibers attract water vapor from the atmosphere as well as any
water present when the paper was produced. Therefore the paper contain cellulose fibers
with a very thin layer of water molecules bound to the surface, and interactions with this
water layer is the most important effect in paper chromatography.
Sample
Sample
Paper
Standard
Lid
Mobile phase front
Container
Starting line
Mobile phase
Figure 17.2 Principle of paper chromatography
264
Introduction to Pharmaceutical Chemical Analysis
With nonpolar mobile phases, paper chromatography can essentially be described as
partitioning chromatography. Polar analytes have strong interactions with the aqueous
stationary phase and are eluted slowly along the paper. More nonpolar substances, in
contrast, have less affinity to the aqueous stationary phase, and they travel with a higher rate
along the paper. When moving to more polar mobile phases, like small alcohols, the
retention mechanism becomes more unclear and is no longer pure partitioning chromatography. Paper chromatography is an official method in the European Pharmacopoeia, but it is
not very much in use.
17.3 Supercritical Fluid Chromatography
In supercritical fluid chromatography, a supercritical fluid is used as mobile phase for the
chromatographic separation. Supercritical CO2 is frequently used for this purpose, and the
term supercritical fluid is illustrated in Figure 17.3.
As illustrated in the figure, carbon dioxide is in a solid state at 78.7 C and 2 bar. As the
pressure is reduced to 1 bar, CO2 is in equilibrium between solid phase and gaseous phase.
This means that CO2 sublimes without turning into a liquid. As the temperature is increased
above the triple point, liquid CO2 is in equilibrium with gaseous CO2 at a certain pressure
following the gas–liquid boundary line. Moving up this line, two phases always exist until
the critical point is reached at 31.3 C. Above this temperature, only one phase exists
regardless of the pressure, and this phase is called a supercritical fluid. Supercritical CO2 is
frequently used as mobile phase in supercritical fluid chromatography (SFC). To ensure
supercritical fluid conditions in the column, the pressure has to exceed 73.9 bar (critical
pressure) and the temperature has to exceed 31.3 C (critical temperature). The temperature
Supercritical
fluid
Pressure (atm)
73.9
Critical point
Solid
Liquid
5
Gas
Triple
point
1
−78.7
−57
31.3
Temperature (ºC)
Figure 17.3 Phase diagram for CO2
Miscellaneous Chemical Techniques
265
of the column is controlled in a column oven similarly to GC to ensure conditions above the
critical temperature. The high pressure in the system is achieved by placing a narrow
capillary at the end of the chromatographic column. This capillary is called a restrictor, and
the strong restriction creates a high pressure in the column. When the supercritical fluid is
leaving the restrictor, it turns to a gas, leaving the solutes in the gas phase for easy detection.
SFC provides increased speed and chromatographic resolution as compared to HPLC.
The reason for this is that diffusion coefficients are higher in supercritical fluids than in
liquids. In contrast, SFC is less efficient than GC because diffusion coefficients are even
higher in gasses than in supercritical fluids. Unlike gases, supercritical fluids can dissolve
nonvolatile analytes. This means that SFC can be used to separate nonvolatile analytes, like
in HPLC. In GC, elution strength is increased by increasing the temperature, and in HPLC
the elution strength is increased by gradient elution. But, in SFC the elution strength is
increased during the chromatographic separation by increasing the density of the supercritical fluid. This can be achieved by gradually increasing the pressure of the supercritical
fluid. In addition, organic modifiers added to the mobile phase, like methanol, can also
increase the elution strength of the mobile phase.
The SFC instrument consists of a pumping system that pumps CO2 into the column at a
pressure above the critical pressure. Injection is typically performed by a loop-injector
similar to that in HPLC. GC-like columns are used for capillary SFC, whereas HPLC-like
columns are used for packed-column SFC. Detectors are usually either flame ionization
detectors or UV detectors. Supercritical fluid chromatography is an official method in the
European Pharmacopoeia, but it is not very much in use.
17.4 Gel Electrophoresis
Gel electrophoresis is a technique used mainly for the separation of macromolecules like
deoxyribonucleic acid (DNA), ribonucleic acid (RNA) and protein molecules. The target
molecules are separated in a gel matrix exposed to a d.c. electrical field. The principle is very
close to capillary electrophoresis, and therefore much of the theory discussed earlier for
capillary electrophoresis also applies to gel electrophoresis. In most cases, the gel used as
separation medium is a crosslinked polymer with a composition and a porosity which is
scaled to the size and nature of the target macromolecules to be separated. For the separation
of proteins or small nucleic acids, the gel is normally composed of acrylamide and a
crosslinker, producing a network of polyacrylamide. For the separation of larger nucleic
acids (greater than a few hundred bases), the preferred matrix is purified agarose. Agarose is
composed of long unbranched chains of uncharged carbohydrate without crosslinks,
providing a gel with large pores that allows for the separation of real macromolecules.
The target molecules to be separated are placed in a well in the gel, and the electrical field
is applied across the gel as illustrated in Figure 17.4.
Upon application of the electrical field, the target molecules begin migrating (moving).
The target molecules migrate towards the anode if they are negatively charged or towards
the cathode if they are positively charged. Different molecules migrate with different
velocities through the gel, largely depending on their molecular size when the charge to
mass ratio of all the molecules is uniform (which is normally the case for DNA and proteins).
The gel is placed in an electrophoresis chamber, which is then connected to a power supply.
266
Introduction to Pharmaceutical Chemical Analysis
Before electrophoresis
During electrophoresis
Wells
-
Molecular size marker
Unknown samples
Migration
direction
Gel / electrophoresis chamber
Power
supply
+
Figure 17.4 Principle of gel electrophoresis
When the electric field is applied, the larger molecules move more slowly through the gel
than the smaller molecules. After a certain period of time, the different target molecules
form distinct bands in the gel depending on their molecular size. To finish the electrophoresis, the electrical field is turned off and the molecules in the gel can be stained to make them
visible (visualization). Ethidium bromide, silver and Coomassie brilliant blue dye are
frequently used for visualization. If the target molecules give fluorescence under UV light, a
photograph can be taken of the gel under a UV lamp.
Several samples can be loaded on the same gel into adjacent wells, and the different
samples are developed in individual lanes. Each lane shows the separation of the components from the original sample as distinct bands, in principle one band per type of
component. However, incomplete separation of the components may result in overlapping
bands. Bands in different lanes that end up at the same distance from the starting point
contain molecules that passed through the gel with the same speed, which usually means
they are approximately of the same size. The position of the different bands can be compared
with a molecular size marker containing a mixture of molecules of known sizes and which
has been loaded into one of the wells in the gel. This enables estimation of the molecular size
of the different separated components. The migration distance of a certain band is
approximately inversely proportional to the logarithm of the size of the molecule.
Gel electrophoresis is used frequently for separation of nucleic acids. For nucleic acids,
the direction of the migration in the gel is from the negative electrode and towards the
positive electrode, due to the naturally occurring negative charge carried by their sugar–phosphate backbone. Double-stranded DNA fragments naturally behave as long rods, so their
migration through the gel is relative to their size or, for cyclic fragments, their radius of
gyration. Single-stranded DNA or RNA tends to fold up into molecules with complex shapes
and migrate through the gel in a more complicated manner based on their tertiary structure.
Therefore, agents that disrupt the hydrogen bonds, such as sodium hydroxide or formamide,
are used to denature the nucleic acids prior to gel electrophoresis and cause them to behave
as long rods.
Miscellaneous Chemical Techniques
267
In contrast to nucleic acids, proteins can have varying charges and complex shapes, and
they are basically less suited for direct gel electrophoresis. Proteins therefore are usually
denatured in the presence of a detergent such as sodium dodecyl sulfate (SDS) or sodium
dodecyl phosphate (SDP) that coats the proteins with a negative charge. Generally, the
amount of SDS bound is proportional to the size of the protein, and usually 1.4 g SDS is
bound per gram of protein. The resulting denatured proteins have an overall negative
charge, and all the proteins have a similar charge to mass ratio. Since denatured proteins
act like long rods instead of having a complex tertiary shape, the rate at which the
resulting SDS-coated proteins migrate in the gel is dependant only on its size and not on
its charge or shape. The latter type of gel electrophoresis is often called sodium dodecyl
sulfate polyacrylamide gel electrophoresis (abbreviated as SDS-PAGE). Gel electrophoresis is an official method in the European Pharmacopoeia and is used primarily for
proteins and peptides.
17.5 Iso-Electric Focusing
Iso-electric focusing (IEF) is a technique for the separation of proteins according to their
iso-electric point (pI). The pI of a protein is the pH value at which the protein carries no net
electrical charge. It is a type of zone electrophoresis which is usually performed in a gel.
IEF takes advantage of the fact that the net charge of a protein changes with the pH of its
surroundings. A protein that is in a pH region below its pI is positively charged and
migrates towards the cathode in an electrical field. As it migrates through a gradient of
increasing pH in the gel, the net charge of the protein decreases until the protein reaches a
pH region that corresponds to the pI of the protein. At this point the protein has no net
charge and the protein no longer migrates. As a result, the proteins become focused into
sharp bands, with each protein positioned at a point in the pH gradient in the gel
corresponding to its pI. This is illustrated in Figure 17.5. The technique is capable of
extremely high resolution, and proteins differing only by a single charge can be separated
into separate bands.
Target molecules to be focused and separated are loaded into a gel with a pH gradient
(usually created by aliphatic ampholytes). An electric current is passed through the
medium, creating an anode (þ) and a cathode () end in the gel. Net negatively charged
proteins migrate through the pH gradient in the medium toward the cathode end, while
positively charged proteins move towards the anode end. As a certain protein migrates in
the gel under the influence of the electrical field, it moves through the changing pH
gradient until it reaches a point in the gel where the pH is equal to the pI of the protein.
At this point the protein no longer has a net electric charge, the positive and negative
charges exactly balance, and as such the protein migrates no further within the gel. The
pH gradient is established before adding the sample of interest to the gel by first
subjecting a solution of small molecules such as polyampholytes with varying pI values
to electrophoresis.
Iso-electric focusing is applied particularly often in the study of proteins. Proteins are
introduced into an immobilized pH gradient gel composed of polyacrylamide, starch, or
agarose, where a pH gradient has been established in advance. Gels with large pores are
usually used in this process to eliminate any “sieving” effects, or artifacts in the pI caused by
268
Introduction to Pharmaceutical Chemical Analysis
Before focusing
After focusing
−
−
pH
pH
10
10
9
9
8
8
7
−
+
−
+
7
Focused bands
6
6
5
5
4
4
3
3
+
+
Figure 17.5 Principle of iso-electric focusing
differing migration rates for proteins of differing sizes. Isoelectric focusing can resolve
proteins that differ in pI value by as little as 0.01. Isoelectric focusing is the first step in twodimensional gel electrophoresis, in which proteins are first separated by their pI and then
further separated by molecular weight through SDS-PAGE. Two-dimensional gel electrophoresis is frequently applied for separation of complex protein samples. Iso-electric
focusing is an official method in the European Pharmacopoeia and is used primarily for
proteins and peptides.
17.6 Nuclear Magnetic Resonance Spectroscopy
Nuclear magnetic resonance spectroscopy, most commonly known as NMR spectroscopy,
is the name given to a technique that exploits the magnetic properties of certain nuclei.
The most important applications related to pharmaceutical analysis are 1 H NMR and
13
C NMR spectroscopy.
Like electrons, the nuclei of certain atoms are considered to spin. The spinning of these
charged particles generates a magnetic moment along the axis of spin, and these nuclei act
like tiny bar magnets. One such nucleus is the proton, the nucleus of ordinary hydrogen 1 H,
and this nucleus is highly important in 1 H NMR spectroscopy.
If a proton is placed in an external magnetic field, the magnetic moment can be aligned in
two ways, either with or against the external magnetic field. Or to say this in another way, the
tiny bar magnet can be aligned either in the direction of the magnetic field, or in the direction
opposite to the magnetic field. Alignment with the magnetic field is more stable than
alignment against the magnetic field. This means that energy must be delivered to the proton
to flip the tiny bar magnet against the magnetic field. The energy required to accomplish this
Miscellaneous Chemical Techniques
269
conversion depends on the strength of the external magnetic field. The stronger the magnetic
field, the greater the tendency for the tiny bar magnet to be aligned with it and the more
energy is needed to flip the bar magnet around. The amount of energy required for this, as
electromagnetic radiation, and expressed in terms of frequency (n) and Hz, is given by this
equation:
n¼
g H0
2
ð17:1Þ
Absorption of radiation
where H0 is the strength of the magnetic field and g is the gyromagnetic ratio, which is
a constant with a value of 26 750 for the proton. In a magnetic field of 14 092 gauss as
an example, the energy required to flip around the spin of a proton corresponds to
electromagnetic radiation of frequency 60 MHz. This is electromagnetic radiation in the
radiofrequency range, and this radiation has much lower energy than infrared light and
visible light.
In 1 H NMR spectroscopy, the substance to be investigated is placed in a magnetic field,
exposed to electromagnetic radiation in the radiofrequency range and the flip around of the
protons in the substance is used to obtain a NMR spectrum. The NMR spectrum gives us
substantial structural information. In principle, the substance can be placed in a magnetic
field of constant strength, and a NMR spectrum can be obtained by passing electromagnetic
radiation (radiofrequency) of a steadily changing frequency through the substance and
observing the radiofrequency at which radiation is absorbed. In practice however, it has been
found more convenient to keep the radiofrequency constant and to vary the strength of the
external magnetic field. At some value of the magnetic field strength, the energy required to
flip a certain proton in the substance matches the energy of the radiofrequency. At this
magnetic field strength, absorption of energy occurs, a signal from this is measured in the
NMR instrument and a peak is observed in the NMR spectrum as illustrated in Figure 17.6.
The key point in 1 H NMR spectroscopy is that the radiofrequency at which a proton
absorbs depends on the magnetic field which that proton feels, and this effective field
Signal
Low field
High field
Magnetic field sweep
Figure 17.6 Simplified NMR spectrum
270
Introduction to Pharmaceutical Chemical Analysis
strength is not exactly the same as the applied field strength. The effective field strength at
each proton depends on the environment of that proton. Among others, this is dependant on
the electron density at the proton and on the presence of other nearby protons and other
electronegative atoms. Each proton, or more precisely, each set of equivalent protons has a
slightly different environment from every other set of protons, and hence it requires a
slightly different applied field strength at which absorption takes place.
At a given radiofrequency, then, all protons absorb at the same effective field strength, but
they absorb at different applied field strengths. It is the applied field strength that is measured
and against which the absorption is plotted. The result is a spectrum showing many
absorption peaks, whose relative positions, reflecting the different environments of the
protons, can give almost unbelievably detailed information about molecular structures.
The number of signals tells us how many different kinds of protons there are in a substance.
The positions of the signals tell us something about the electronic environment of each kind
of proton. The intensity of the signals tells us how many of each kind of proton there are.
Finally, the splitting of a signal into several peaks tells us about the environment of a proton
with respect to other nearby protons. NMR spectroscopy is an official method in the
European Pharmacopoeia and is primarily used for identification purposes, but it is also used
for a few quantitative applications. Many types of information can be obtained from a NMR
spectrum. Much like using infrared spectroscopy (IR) to identify functional groups, analysis
of a NMR spectrum provides information on the number and type of chemical entities in a
molecule. However, NMR provides much more information than IR.
17.7 Raman Spectrometry
Raman spectroscopy is used for identification purposes and can be considered to be
complementary to IR spectroscopy. When electromagnetic radiation passes through a
transparent medium, the species present scatter a small fraction of the beam in all directions.
The majority of the scattered light has the same wavelength as the incident beam, and this is
called Rayleigh scattering. In addition to the Rayleigh scattering, a small fraction of the
radiation scattered by certain molecules differs from that of the incident beam, and this is
called Raman scattering. Raman scattering is observed as radiation bands at different
wavelengths. The shifts in wavelength, namely the difference between Raman scattering
bands and the Rayleigh scattering, depend upon the chemical structure responsible for the
scattering. These shifts are called Raman shifts and are the basis for Raman spectroscopy.
Raman spectra are obtained by irradiation of a sample with a powerful laser source of
visible or infrared monochromatic radiation. During irradiation, the spectrum of the
scattered radiation is measured at some angle (usually 90 ) with a suitable spectrometer.
An example of a Raman spectrum for CCl4 is shown in Figure 17.7.
This spectrum was obtained by irradiation with an intense beam of an argon laser having a
wavelength of 488.0 nm. The major band measured for the scattered radiation is the
Rayleigh scattering at 488.0 nm, corresponding to a wavenumber of 20 492 cm1, exactly
at the same wavelength as the excitation source. Several bands are also observed at lower
wavenumbers (lower frequency) and these bands, which are structure-specific for CCl4, are
termed Stoke lines. Additionally, some bands are observed at higher frequencies than the
Rayleigh scattering, and these bands are termed anti-Stoke lines.
Miscellaneous Chemical Techniques
271
Rayleigh scattering
Stokes lines
Anti-Stokes lines
Absorbance
−459
−314
−218
218
−400
−200
0
314
200
459
400
Wavenumber (cm −1)
Figure 17.7 Raman spectrum for CCl4
In a Raman spectrum, the abscissa is the wavenumber shift (D
n), which is defined as the
difference in wavenumber (cm1) between the observed radiation and that of the source. For
CCl4, three Raman peaks are found on either side of the Rayleigh peak and the patterns of the
shifts on the two sides are identical. That is, Stoke lines are found at wavenumbers that are
218, 314 and 459 cm1 smaller than the Rayleigh peak, whereas anti-Stoke lines occur at
218, 314 and 459 cm1 greater than the wavenumber of the source. It is important to notice
that the magnitude of the Raman shifts is independent of the wavelength of excitation. Thus,
shift patterns identical to those shown in Figure 17.7 would be observed for CCl4 regardless
of what type of laser is used for the excitation.
The molecular phenomena responsible for the Raman scattering are of the same type of
those responsible for infrared absorption. Thus, the difference in wavelength between the
incident and the scattered radiation corresponds to wavelengths in the mid-infrared region.
Indeed, the Raman scattering spectrum and infrared absorption spectrum for a given species
often resemble one another quite closely. There are, however, enough differences between
the kinds of groups that are infrared active and those that are Raman active to make the
techniques complementary rather than competitive. For some problems, the infrared
method is the superior tool; for others, the Raman procedure offers more useful spectra.
An important advantage of Raman spectra over infrared lies in the fact that water does not
cause interferences. Thus, Raman spectra can be obtained from aqueous solutions. In
addition, glass and quartz cells can be employed in Raman spectroscopy, thus avoiding the
inconvenience of making potassium bromide tablets of the substances to be investigated.
Raman spectroscopy is an official method in the European Pharmacopoeia and is used for
identification purposes.
18
Sample Preparation
Many samples of pharmaceutical interest have a very complex composition, and they cannot
be analyzed directly by chromatographic or spectroscopic methods. Sample preparation is
required to reduce the complexity of the sample and to make it compatible with the
chromatographic or spectroscopic method. This chapter discusses different strategies for
sample preparation. Important terms and parameters used in connection with sample
preparation will also be defined and discussed. Applications of sample preparation in
pharmaceutical analysis are mainly discussed in Chapters 22 and 23.
18.1 Why is Sample Preparation Required?
Many samples are complex mixtures of different ingredients. In pharmaceutical analysis the
samples are typically finished pharmaceutical products like tablets or ointments, or
biological fluids like blood and urine. In finished pharmaceutical products, the active
pharmaceutical ingredient (API) is normally mixed with various excipients, and this
complicates the identification and the quantitative analysis of the API. Blood and urine
are also very complex mixtures consisting of hundreds or thousands of components, and the
drug substance of interest (analyte) is present in a very low concentration. Low concentration in this context is defined from high pg/ml to low mg/ml level.
Analysis of drugs in complex mixtures is challenging for several reasons:
1. Complex mixtures can contain one or more substances that provide a false response
(interference) during the analytical measurement.
2. Complex mixtures can contain substances that contaminate or even destroy the analytical
instrument.
3. The concentration of the analyte can be so low that it cannot be detected by the analytical
instrument.
4. Complex mixtures (samples) can be incompatible with the analytical instrument.
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
274
Introduction to Pharmaceutical Chemical Analysis
Challenges 1, 2 and 4 are typical for the quality control of finished pharmaceutical products.
Some of the excipients are responding to the analytical measurement, and the response or
analytical signal from these cannot be separated from the response from the API. Thus, the
interfering excipients give a false positive contribution to the analytical measurement, and
this results in serious errors both for identification and for quantitative analysis. Many
finished pharmaceutical products are not compatible with analytical instruments because
they are solid or semisolid. One example is tablets, and these have to be dissolved to be
compatible with most analytical instruments. All of challenges 1, 2, 3 and 4 are typical for
the analysis of drug substances in biological samples. Biological samples have a complex
composition, and there is always a risk that some of the matrix components can give a
response (interference) that cannot be separated from the analyte. Several substances in
biological samples can also damage the analytical instrument, and one example of this is the
presence of proteins in blood samples that can clog HPLC columns. The concentration of
many drugs is often too low for direct detection, as they are present below the limit of
quantification or limit of detection. The solution to all of these challenges is to do sample
preparation. Which strategy is to be used is determined by the properties of the analyte, by
the sample composition, and by the concentration of the analyte in the sample.
18.2 Main Strategies
This chapter discusses four different approaches for sample preparation in the pharmaceutical laboratory:
.
Protein precipitation;
Liquid–liquid extraction;
. Solid–liquid extraction;
. Solid-phase extraction.
.
The principle for protein precipitation (PP) is illustrated in Figure 18.1. PP is used for
plasma and serum samples that contain proteins, and the main purpose of PP is to get rid of
the proteins prior to HPLC. This is important because proteins can clog the HPLC column.
The plasma or the serum sample is mixed with acetonitrile, which causes the proteins to
precipitate. The sample is centrifuged, and the supernatant is collected for the final analysis.
The supernatant contains the analyte of interest, but is almost free of proteins.
PP is discussed in more detail in Section 18.4. A second technique frequently used for
sample preparation is liquid–liquid extraction (LLE). This technique is used for biological
Plasma sample
Add aceto-
Shake and
nitrile
centrifuge
Supernatant for analysis
Protein precipitate
Analyte molecules
Figure 18.1 Principle of protein precipitation
Sample Preparation
Aqueous sample
Add organic
Shake and
solvent
centrifuge
275
Extract for analysis
Analyte molecules
Figure 18.2 Principle of liquid–liquid extraction
fluids, like blood and urine, and also for many liquid pharmaceutical products such as oral
solutions and ointments. The sample solution is in most cases aqueous. The principle for
LLE is illustrated in Figure 18.2. The aqueous sample is mixed with an organic solvent
immiscible with water, resulting in a two-phase system. This two-phase system is shaked
vigorously for a certain period of time, and small droplets of the organic phase mix with the
water phase. During this shaking, the analytes are transferred from the aqueous sample and
to the organic phase, which is called the extract. After shaking, the mixture is centrifuged,
and the two phases separate again. The extract (organic phase) containing the analyte is then
collected for the final analysis. LLE is discussed in more detail in Section 18.5.
In the case of solid samples, sample preparation can be accomplished by solid–liquid
extraction (SLE). This is typically the case when analyzing tablets or capsules. Figure 18.3
illustrates the principle for SLE. The solid sample to be analyzed is powdered and an organic
solvent or an aqueous solvent is added. The mixture is shaked for certain period of time, and
the analyte is dissolved into the solvent. The solvent containing the analyte (extract) is them
collected for the final analysis. SLE is discussed in more detail in Section 18.6.
The final sample preparation method discussed in this textbook is solid-phase extraction (SPE). The principle for SPE is illustrated in Figure 18.4. In SPE a small column
packed with a stationary phase is used, which typically resembles the stationary phases
discussed earlier for HPLC. The sample, which normally is a biological fluid like blood or
urine, is drawn through the column in step 1. This is called loading, and the analytes of
interest are retained in the column by different type of interactions with the stationary
phase. In step 2, which is called washing, the column is washed to remove as many matrix
components as possible, but without washing the analytes from the column. Finally, in
step 3, which is called elution, the analytes are eluted from the column with a solvent that
break the analyte–stationary phase interactions, and this eluate (extract) is collected for
the final analysis.
SPE is discussed in more detail in Section 18.7.
Analyte molecules
Extract for analysis
Powder
Transfer to
Shake and
solid sample
solvent
centrifuge
Sample matrix
Figure 18.3 Principle of solid–liquid extraction
276
Introduction to Pharmaceutical Chemical Analysis
Loading of sample
Analyte
molecules
Column
Matrix
components
Stationary
phase
Washing
Waste
Step 1
Step 2
Elution
Eluate for analysis
Step 3
Figure 18.4 Principle of solid–phase extraction
18.3 Recovery and Enrichment
Before going more into the different techniques, the terms recovery and enrichment have to
be defined, which are important parameters to characterize the performance and success of a
sample preparation method.
Recovery (R) is given as a percentage and is defined by the following equation:
nE;final
R¼
100%
ð18:1Þ
nS;initial
where nE;final is the amount of analyte collected in the extract ready for the final analysis
and nS;initial is the amount of analyte present in the sample before the sample preparation.
Recovery gives the percentage of analyte collected during the sample preparation, and
when R is 100% this means that the total amount of analyte is transferred from the sample
and into the extract. When R is 25% it means that one-quarter of the analyte molecules has
been transferred into the extract. Normally, it is desirable that recoveries are as close to
100% as possible. For practical applications, the following equation can be used to
calculate recoveries:
R¼
CE;final VE
100%
CS;initial VS
ð18:2Þ
where CE;final is the concentration of analyte finally present in the extract, VE is the volume
of the extract, CS;initial is the concentration of analyte initially present in the sample and VS
is the volume of the sample.
Recoveries are used to characterize a sample preparation method. For quantitative
analytical methods, where several analytes are measured simultaneously, the method
should specify the exact recovery for each of the analytes because different analytes
can be extracted with different efficiencies. Recoveries are measured during method
development and validation. In addition, if the recovery for a certain analyte is less than
100%, the recovery has to be measured every time the method is calibrated with standards
for quantitative analysis. This is important in order to correct the analytical results for the
fact that the extraction is not complete. This is discussed in more detail below. Box 18.1
illustrates an example on how to determine the recovery for an extraction method in real life.
Sample Preparation
277
Box 18.1 Determination of extraction recovery
A new drug is extracted from plasma by LLE and the extract is analyzed by HPLC.
Because the method is new, the recovery for the LLE procedure has to be established.
LLE is accomplished as follows:
1 ml of plasma is extracted with 4 ml of an organic solvent. After extraction and
separation of the two phases, the organic phase is taken away and evaporated to dryness.
The residual is dissolved in 0.25 ml of mobile phase and 50 ml is injected into the HPLC.
The peak area for the drug is recorded.
To measure the recovery the following solutions are prepared:
Solution 1: The drug spiked into a drug-free plasma sample to give a drug concentration
of 1.0 mg/ml.
Solution 2: The drug dissolved into mobile phase to give a drug concentration of
4.0 mg/ml.
Solution 1 is extracted according to the procedure, and the extract is analyzed by HPLC
giving a peak area of 11 263. Solution 2 is analyzed directly by HPLC, giving a peak area
of 12 467. The concentration of solution 2 is 4.0 mg/ml, and from this the correlation
between peak area and concentration can be established (12 467 in peak area corresponds
to 4.0 mg/ml in concentration). First the concentration of drug in solution 1 is calculated,
which equals CE;final in Equation (18.2):
CE;final ¼
11 263
4:0 mg=ml ¼ 3:6 mg=ml
12 467
Then the recovery is calculated based on Equation (18.2):
R¼
CE;final VE
3:6 mg=ml 0:25 ml
100% ¼ 90%
100% ¼
1:0 mg=ml 1:00 ml
CS;initial VS
Enrichment (E ) is defined by the following equation:
E¼
CE;final
VS R
¼
CS;initial VE 100%
ð18:3Þ
Enrichment tells how many times the analyte has been pre-concentrated during the sample
preparation. Enrichment is important when analyzing drugs at very low concentration levels
(pg/ml or low ng/ml level), to ensure that the extract contains a concentration which is above
the limit of quantification for the analytical method. Enrichment is most relevant for analysis
of biological fluids, and under certain conditions, enrichments up to 10–20 times can be
achieved. Enrichment is normally not required for the analysis of finished pharmaceutical
products because API concentrations are relatively high. Box 18.2 illustrates an example on
how to calculate enrichment.
278
Introduction to Pharmaceutical Chemical Analysis
Box 18.2 Determination of enrichment
In Box 18.1 the recovery was calculated for a new extraction method. Based on the data
from this experiment, the enrichment is calculated based on Equation (18.3):
E¼
CE;final
3:61 mg=ml
¼ 3:61
¼
CS;initial 1:00 mg=ml
This value means that the concentration of drug in the final extract was 3.61 times higher
than in the original sample.
18.4 Protein Precipitation
Protein precipitation (PP) is a very popular sample preparation prior to HPLC and LC-MS
analysis of biological fluids, and the reason for this is that it is a very simple procedure. PP is
used when samples contain large amounts of proteins, and this is the case with plasma and
serum samples. These samples contain about 8% (w/w) of protein. Urine samples do not
contain large amounts of proteins, and they are not prepared by PP. The purpose of PP is to
get rid of the proteins. This is very important because proteins can adsorb to HPLC columns
and reduce their performance.
As mentioned in Section 18.2, the typical PP procedure involves pipetting a certain
volume of plasma or serum, and this sample is then mixed with a certain volume of
acetonitrile. Typically, the volume of sample is between 0.1 and 1.0 ml. PP is performed in
small centrifuge tubes. Acetonitrile is miscible with plasma or serum, making a single-phase
system, but the presence of acetonitrile in the sample reduces the solubility of the proteins
and they will precipitate. Acetonitrile therefore acts as a precipitant. The mixture is shaked
vigorously for a short period of time to facilitate the precipitation, and subsequently the
mixture is centrifuged. After centrifugation, the clear supernatant is collected for direct
injection in HPLC/LC-MS.
PP is definitely a sample preparation method, but strictly speaking it is not an extraction
method as the analytes do not transfer to another phase during the procedure. The other
sample preparation methods discussed later (LLE, SLE, SPE) are true extraction methods.
PP can also be performed with other organic solvents as illustrated in Table 18.1, among
others acetone and methanol.
As seen from Table 18.1, acetone and methanol are less efficient precipitants than
acetonitrile. It is also clear from Table 18.1 that the amount of precipitant plays an
important role, as the efficiency of the protein removal increases with the volume of
precipitant. With 1.0 ml plasma as an example, 0.2 ml acetonitrile gives a removal
efficiency of only 13%, whereas this improves to about 97% as the volume of acetonitrile
is increased to 1.0 ml. If the volume is increased to 2.0 ml, the protein removal is almost
complete (99.7%). As illustrated in Table 18.1 different acids, like trichloroacetic acid or
perchloric acid, and different metal ions can also be used as precipitants. This enables PP
procedures to be carefully optimized. However, acetonitrile is by far the most frequently
used precipitant in PP.
Sample Preparation
279
Table 18.1 Different approaches to protein precipitation
Precipitant
Acetonitrile
Acetone
Methanol
CCl3COOH (10%)
HClO4 (6%)
CuSO4–Na2WO4
ZnSO4–NaOH
Volume of precipitant added per volume of plasma
0.2
1.0
Protein precipitation efficiency (%)
2.0
13.4
1.5
17.6
99.7
35.4
36.5
41.1
97.2
96.2
73.4
99.5
99.1
97.5
94.2
99.7
99.4
98.7
99.8
99.1
99.9
99.3
18.5 Liquid–Liquid Extraction
18.5.1
Fundamentals
Liquid–liquid extraction (LLE) is a popular technique for sample preparation of
biological fluids (blood, urine) and it is also used for many aqueous pharmaceutical
products (like oral solutions). For LLE of biological fluids, the sample volume is limited
and the typical LLE procedure starts with pipetting a small volume of the aqueous sample
into a centrifuge tube. Especially for plasma and serum, it is desirable to use less than
1 ml of biological fluid. For LLE of finished pharmaceutical products, larger sample
volumes are available, typically in the range 10–100 ml, and in such cases LLE is
typically upscaled somewhat and performed in a separation funnel. Separation funnels
are known from the organic chemistry laboratory. Also, as discussed in more detail later,
the pH is adjusted in the sample to ensure an efficient extraction of the analyte. This can
be done by adding a buffer solution to the sample, or solutions of a strong acid or base.
Further, the extraction solvent is added to the sample. It is important that the solvent is
immiscible with water to form a two-phase system with the sample. Then, the centrifuge
tube or the separation funnel is shaked for a certain period of time to ensure efficient
mixing of the two phases. During this, the analyte is transferred from the aqueous sample
and into the small droplets of organic solvent formed by shaking. After this, the
centrifuge tube or the separation funnel is allowed to stand for the two phases to separate
again. Especially with biological fluids, this phase separation can be difficult, but it is
promoted by centrifugation of the mixture. After centrifugation, the organic phase can be
collected as the extract containing the analyte, and basically it is ready for the final
analytical measurement.
18.5.2
A Closer Look at the Theory
From a more fundamental point of view, the LLE process can be described in the
following way:
AAqueous $ AOrganic
ð18:4Þ
280
Introduction to Pharmaceutical Chemical Analysis
where AAqueous represents the analyte present in the aqueous sample and AOrganic represents
the analyte present in the organic extract. The analyte is distributed between the two
phases during LLE, and this distribution can be described by the partition ratio KD:
AOrganic
KD ¼ ð18:5Þ
AAqueous
where AOrganic is the concentration of the analyte in the organic phase and AAqueous is
the concentration of analyte in the aqueous phase at equilibrium. KD is constant for a given
analyte under given conditions in the aqueous and organic phases. The higher the partition
ratio, the more efficiently the analyte is extracted into the organic phase.
For neutral analytes, Equation (18.5) describes the situation well. However, many
pharmaceutical compounds are either acidic or basic compounds, and for these the situation
becomes a little more complicated because they dissociate in aqueous solution. Considering
extraction of an acidic analyte, which dissociate in aqueous solution according to the
following scheme:
HA $ H þ þ A
ð18:6Þ
and with the following acid dissociation constant:
Ka ¼
½H þ ½A ½HA
ð18:7Þ
For this compound the distribution ratio, D, can be defined according to the following
equation:
HAOrganic;total
D¼
ð18:8Þ
HAAqueous;total
is the total concentration of the analyte in the organic phase and
where HAOrganic;total
HAAqueous;total is the total concentration of analyte in the aqueous phase. The key point
is that only the uncharged fraction of the analyte can be extracted from the aqueous
sample and into the organic phase, whereas the charged species will remain in
the
aqueous
sample.
Thus, the following equations can be written for HAOrganic;total and
HAAqueous;total respectively:
ð18:9Þ
HAOrganic;total ¼ HAOrganic
i
h HAAqueous;total ¼ HAAqueous þ AAqueous
ð18:10Þ
By substitution of Equations (18.9) and (18.10) into (18.8) one gets the following:
HAOrganic
h
i
D¼
ð18:11Þ
HAAqueous þ AAqueous
By substitution of Equations (18.5) and (18.7) into (18.11) one gets:
D ¼ KD ½H þ ½H þ þ Ka
ð18:12Þ
Sample Preparation
281
In a similar way, the following equation can be obtained for extraction of a basic analyte A
which is partly dissociated in the aqueous sample:
A þ H þ $ AH þ
D ¼ KD ð18:13Þ
Ka
½H þ þ K a
ð18:14Þ
where Ka is the dissociation constant for the corresponding acid AH þ . In practical LLE,
one tries to get as high a recovery as possible, and to obtain this, one tries to maximize the
values for the distribution ratio (D). Equations (18.12) and (18.14) show that the following
parameters affect D and therefore affect the extraction recovery:
.
Partition ratio;
pH in the aqueous sample;
. pKa value of the analyte.
.
For a neutral analyte, with no acidic or basic functional groups, the extraction recovery is
independent of pH in the aqueous sample, and no pH adjustment is required. For neutral
analytes, recovery is determined by the partition ratio only. A high partition ratio gives a
high recovery. For acidic and basic analytes, pH in the sample plays a major role, and pH
adjustment in the sample is normally required prior to extraction. For efficient extraction of
acids, Equation (18.12) shows that pH should be low by maintaining acidic conditions in the
sample. Under acidic conditions, the distribution ratio is very close to the partition ratio.
Figure 18.5 shows how the distribution ratio varies as function of pH. From this figure it is
clear that the highest distribution into the organic phase is obtained at pH values at least two
units below the pKa value for the acidic analyte. For highest distribution into the aqueous
phase in contrast, the pH should be high and preferably two units higher that the pKa value.
Figure 18.6 shows a similar plot for a basic analyte.
In this case, highest distribution into the organic phase is obtained at a pH at least two
units higher than the pKa value. In other words, to effectively extract basic compounds into
organic solvents, alkaline conditions should be used in the sample. By using alkaline
conditions, the ionization of the analyte is suppressed, and therefore extraction into the
organic solvent is promoted.
4.2
log D
3.4
Ibuprofen
2.6
1.9
1.1
0.3
0
2
4
6
8
10
12
14
pH
Figure 18.5 Distribution ratio as a function of pH for an acidic analyte
282
Introduction to Pharmaceutical Chemical Analysis
5.1
Amitriptyline
log D
4.3
3.6
2.9
2.2
1.5
0
2
4
6
8
10
12
14
pH
Figure 18.6 Distribution ratio as a function of pH for a basic analyte
18.5.3
Extraction Solvents
In addition to pH in the sample, the organic solvent is very important in LLE. To obtain high
partition ratios and therefore high extraction recoveries, a good solvent for the particular
analyte should be selected. For very nonpolar analytes, which are hydrocarbon-like
substances with little functional groups, the best solvents are the nonpolar solvents. Typical
extraction solvents are aliphatic hydrocarbons like n-heptane or cyclohexane as mentioned
in Table 18.2. For analytes of low polarity, but with aromatic character, aromatic solvents
like toluene or p-xylene can be more efficient. For the more polar analytes, a more polar
solvent should be selected, but still not too polar because the solvent should be immiscible
with water to form a two-phase system. For polar analytes, solvents with high dipole
moments are very efficient. Examples of solvents with good dipole properties are dichloromethane and ethyl acetate. For the extraction of basic analytes, which are characterized by
their tendency to accept protons and which are normally amines, it can be a good idea to use
Table 18.2 Some examples of LLE solvents and their properties
Solvent
Non-polar solvents
n-Hexanea
Cyclohexane
n-Heptane
Aromatic solvents
Toluene
p-Xylene
Dipolar solvents
Dichloromethane
Ethyl acetate
Proton-donating solvents
Chloroforma
Proton-accepting solvents
Methyl-tertbutyl ether
Diethyl ether
a
Polarity
index
Boiling
point ( C)
Viscosity
(cPoise)
Water solubility
(% w/w)
Density
(g/ml)
0.1
0.2
0.1
69
81
98
0.30
1.0
0.42
0.014
<0.1
0.01
0.655
0.779
0.684
2.4
2.5
111
138
0.59
0.81
0.05
Insoluble
0.867
0.861
3.1
4.4
40
77
0.44
0.43
0.9
8.7
1.327
0.894
4.1
61
0.53
0.8
1.498
2.5
2.9
55
35
0.27
0.24
5.1
7.5
0.741
0.713
Should be avoided if possible due to toxicity.
Sample Preparation
283
solvents with proton-donating properties. Among the most prominent solvents with protondonating properties is chloroform, but several other solvents also has acceptable protondonating properties. For acidic analytes, which are characterized by their tendency to donate
protons, proton-accepting solvents should preferably be used. Typical examples of protonaccepting solvents are diethyl ether and methyl-tertbutyl ether.
As discussed above, polar analytes have to be extracted with relatively polar organic
solvents, and nonpolar analytes have to be extracted with the more nonpolar solvents. The
polarity of solvents can actually be quantified by the polarity index as illustrated in
Table 18.2. For the most non-polar aliphatic hydrocarbons, the polarity index is very
close to 0. As the polarity increases, the polarity index also increases, and water has a
polarity index of 10.2. Solvents with polarity index below 4 or 5 are normally immiscible
with water and make a two-phase system with aqueous samples, whereas solvents with a
polarity index above 5 are totally miscible with water and can not be used for LLE.
The choice of solvent for LLE does not solely focus on analyte solubility. Other properties
of the solvents are also crucial. The density and viscosity are important for the practical
implementation of the extraction method. Solvents with low viscosity are preferred because
they more easily brake up into small droplets and mix with the aqueous sample during
extraction, and this is beneficial for the transfer of analyte into the organic phase. Solvents
with a lower density than water will float on top of the aqueous sample after extraction. This
is an advantage if the organic extraction is to be collected by a pipette after extraction from a
small centrifuge tube. Examples of solvents less dense than water are hexane and toluene. If
a solvent more dense that water is used, it will be recovered below the aqueous sample after
LLE. Examples of solvents with higher density than water are chloroform and dichloromethane. The volatility of the solvent and the solubility in water are also crucial. Any
evaporation of solvent after extraction is much faster with a solvent with low boiling point,
and volatile solvents are normally preferred if they have to be evaporated after extraction.
Finally, for the solvent selected for extraction, the solubility in water should be low in order
to avoid that substantial amounts of the extracting solvent is dissolved into the aqueous
sample. As seen in Table 18.2, the water solubility of solvents can be quite different; the
nonpolar solvent hexane shows neglectible water solubility, whereas ethyl acetate dissolves
into water in substantial amounts.
18.5.4
Calculation of Recovery
Equations (18.12) and (18.14) can be used to calculate the distribution ratio for LLE
applications. These equations also demonstrates that the distribution ratio is very close to the
partition ratio if the extraction is carried out at optimal pH (low pH for acids and high pH for
bases). Under optimal pH conditions, the partition ratio can be used to calculate the expected
recovery from LLE, but how? Suppose that analyte A is extracted under optimal pH
conditions from VAqueous to VOrganic with the partition ratio KD and with the recovery R.
Initially, m moles of A were present in the sample. After the LLE has finished, the organic
phase contain R m moles of A, whereas the sample contain (100 R) m moles of A.
This can be inserted into the equation for the partition ratio:
Rm
AOrganic
V
¼ ð100Organic
KD ¼ RÞ m
AAqueous
VAqueous
ð18:15Þ
284
Introduction to Pharmaceutical Chemical Analysis
Box 18.3 Calculation of theoretical recoveries
1.0 ml plasma was extracted with 1.0 ml organic solvent, and the partition ratio for the
analyte in this particular system was 10. The recovery is calculated based on the
following equation:
10 1:0
100% ¼ 91%
R¼
10 1:0 þ 1:0
This recovery is not satisfactory. A similar extraction is conducted where the volume of
organic solvent is increased to 5.0 ml:
R¼
10 5:0
100% ¼ 98%
10 5:0 þ 1:0
Clearly, the recovery in LLE can be increased by increasing the volume of organic
solvent. Alternatively one can try to optimize the organic solvent to increase the partition
ratio. Suppose that the initial extraction is carried out with a solvent providing a partition
ratio of 30:
30 1:0
100% ¼ 97%
R¼
30 1:0 þ 1:0
Clearly, one can also increase the recovery in LLE by increasing the partition ratio. If
both approaches are combined, increasing the volume of organic solvent and the
partition ratio, one get:
R¼
30 5:0
100% ¼ 99%
30 5:0 þ 1:0
In the latter case, the extraction is essentially quantitative or exhaustive, which means
that one extracts the total amount of analyte in the sample.
Rearrangement of Equation (18.15) gives the following equation:
R¼
KD VOrganic
100%
KD VOrganic þ VAqueous
ð18:16Þ
Examples on how to use this equation to calculate expected recoveries are shown in
Box 18.3.
From Equation (18.16) the expected recovery from any LLE of a certain analyte can be
calculated when the volumes of sample and the organic phase have been decided, provided
that the partition ratio is known. But how can information about the partition ratio be
obtained? Partition ratios vary from compound to compound, and are very dependant on the
experimental conditions. Especially, partition ratios are very dependant on the type of
solvent used for LLE as discussed above, but they also vary with temperature and with the
presence of matrix components in the aqueous sample. Therefore, partition ratios can not
be found in general in the literature. Nevertheless, some valuable information can be
obtained from the literature from tables with log P values. As discussed earlier in this book
(Chapter 3), the log P value for a certain compound gives the logarithmic value for the
Sample Preparation
285
partition ratio between 1-octanol and water. Theoretically, recoveries using 1-octanol as the
extraction solvent can be calculated. However, 1-octanol is not used to any extent as
extraction solvent in LLE, among others because it has a high boiling point and because it is
incompatible with most analytical instruments. However, a high log P value indicates that
the compound is relatively non-polar, and the non-polar analytes are the easiest to extract by
LLE because they have high partition ratios in general. As a rule of thumb, compounds with
log P values below 0 are very polar and difficult to extract with any solvent in LLE.
Compounds with log P between 0 and 1 are relatively polar, and they are not the best
candidates for LLE. Compounds with log P values above 1 are more hydrophobic, and they
are generally well extracted in LLE.
18.5.5
Multiple Extractions
One prefers to extract 100% of the analyte during LLE. In some cases, this can be
accomplished in a single extraction as discussed in Box 18.3. The requirement for this
is that the partition ratio is relatively large. If not, the volume of the organic phase can be
increased as illustrated in Box 18.3 to compensate for the poor partition ratio. However,
another way of increasing the recovery, which is even more effective, is to extract the
aqueous sample with several fresh portions of organic solvent, and after that the different
portions of organic solvent are combined. This is called multiple extractions which was
exemplified in Example 3.2 (Chapter 3). Suppose that analyte A is extracted under optimal
pH conditions from VAqueous to VOrganic with the partition ratio KD and where the fraction of
A remaining in the sample after LLE is q. Initially, m moles of A was present in the sample.
After the LLE has been finished, the sample contain q m moles of A, whereas the organic
phase contain (1 q) m moles of A. This can be inserted into the equation for the
partition ratio:
ð1 qÞ m
AOrganic
V
¼ Organic
KD ¼ ð18:17Þ
qm
AAqueous
VAquoeus
This equation can be rearranged to the following equation describing the amount (fraction)
of analyte left in the sample after LLE, which states the amount that has NOT been extracted:
Unextracted ¼ q ¼
VAqueous
VAqueous þ KD VOrganic
ð18:18Þ
If the organic phase is removed and a new portion of organic solvent is added to the sample
for a second LLE step, the fraction of analyte remaining in the sample after two extractions is
given by:
2
VAqueous
Unextracted ¼ q q ¼
ð18:19Þ
VAqueous þ KD VOrganic
After n extractions of the sample, the generalized equation describing how much analyte is
left in the sample is as follows:
n
VAqueous
Unextracted ¼ qn ¼
ð18:20Þ
VAqueous þ KD VOrganic
286
Introduction to Pharmaceutical Chemical Analysis
Box 18.4 Calculation of theoretical recovery for multiple extractions
In Box 18.3 the recovery for a certain analyte from a single extraction of 1.0 ml of plasma
with 1.0 ml of organic solvent was calculated and with a partition ratio of 10. This gave a
91% recovery.
Suppose that one extracts the same sample with two portions of 0.5 ml each as an
alternative procedure, and the recovery can be calculated as follows:
2
1:0
Unextracted ¼
¼ 0:028
1:0 þ 10 0:5
This means that the recovery after two extractions is:
R ¼ ð1 0:028Þ 100% ¼ 97%
If a third extraction is performed with 0.5 ml organic solvent and all three portions of
solvent are combined one gets:
0
13
1:0
A ¼ 0:005
Unextracted ¼ @
1:0 þ 10 0:5
R ¼ ð1 0:005Þ 100% ¼ 99:5%
These calculations clearly demonstrate the effectiveness of multiple extractions.
Box 18.4 illustrates how Equation (18.20) can be used to calculate recoveries from multiple
extractions.
18.5.6
LLE with Back-Extraction
LLE as discussed so far has been based on the fact that the analytes are transferred from the
aqueous sample and into an organic solvent (single extraction). Subsequently, the organic
solvent or the organic extract is directed to the final chemical analysis. With very
complicated samples, containing many matrix components at high concentration, such
a single extraction may not be sufficient as some of the matrix compounds also can be
extracted into the organic solvent. These matrix components can be different biological
substances in biological fluids, or it can be excipients of low water solubility from
pharmaceutical preparation. In such cases, it can be an advantage to do a second extraction
step, called back-extraction, into a new aqueous phase which is used for the final chemical
analysis. Thus, the organic extract from the first extraction is extracted itself with a new
aqueous phase in a second step. The advantages of back-extraction are both improved
clean-up of the sample, and also that the final extract now is aqueous, and more compatible
with certain analytical instruments. A typical scheme for extraction followed by backextraction is as follows; a basic API is extracted from an aqueous sample by adjusting pH
in the sample into the high alkaline region, and into an organic solvent. Subsequently, the
organic extract is collected into a new extraction tube, and an acidic aqueous solution is
added as a second extraction fluid. Because basic compounds normally have high
Sample Preparation
287
solubility in acidic solution, the basic analyte is back-extracted from the organic solvent
and into the acidic aqueous extract.
18.6 Solid–Liquid Extraction
Solid–liquid extraction is abbreviated SLE, and SLE is used to extract analytes from solid
samples by an extracting solvent. In the pharmaceutical laboratory, SLE is mostly used for
extraction of APIs from solid pharmaceutical products, like tablets and capsules. This will
be discussed in more detail in Chapter 22. SLE is also an extraction method, similar to LLE,
because the analyte is transferred from one phase to another; in this case from a solid sample
and into a liquid. The theoretical considerations discussed in LLE also apply to SLE and
should not be repeated in this section.
The typical SLE procedure starts with milling or powdering of the solid sample to be
investigated. This is important to improve the extraction efficiency or the recovery. As the
sample is disintegrated into small particles, the analyte molecules included in the solid
sample becomes much more accessible to the solvent. After disintegration of the sample, a
certain amount of sample is transferred to the extraction vessel. Then the extraction solvent
is added, and the whole system is shaked, vibrated, or treated in an ultrasonification bath for
a certain period of time. Shaking or ultrasonification is important to speed up the extraction.
During the extraction, the analyte molecules are released from the solid sample and
dissolved into the extraction solvent. After extraction, the content of the vessel is filtrated
in order to separate the residuals of the solid sample from the extraction solvent. The filtrate
is then used for the final chemical analysis.
SLE differs from LLE in two different ways. First, it is more difficult to extract analytes
from a solid sample, and normally, SLE is performed as multiple extractions, where the
same sample is extracted several times with new portions of solvent, and where the solvent
portions afterwards are combined to the final extract. Second, in SLE one can also use
solvents miscible with water, as long as the major matrix constituents are not dissolved by
the solvent. This means that methanol, ethanol, and acetonitrile can be used as extraction
solvents in SLE. Actually, water itself can be used as extraction solvent, or more
frequently, aqueous solutions of a strong base or a strong acid. Aqueous solutions of
strong bases are used to extract acidic analytes, whereas aqueous solutions of strong acids
are used to extract basic analytes. Recall that basic compounds have their highest water
solubility under acidic conditions.
18.7 Solid Phase Extraction
18.7.1
Fundamentals
Solid phase extraction is abbreviated SPE, and is frequently used for sample preparation of
biological fluids like plasma, serum and urine. It is mostly used for extraction of analytes
from aqueous sample solutions, but can also be used for liquid samples that are nonaqueous. As briefly discussed in Section 18.2, SPE utilizes a small extraction column as
illustrated in Figure 18.7, packed with a certain stationary phase. The principles of SPE were
288
Introduction to Pharmaceutical Chemical Analysis
Reservoir
SPE column
Stationary phase
Filters
Figure 18.7 Solid-phase extraction column
illustrated in Figure 18.4. The first step in the procedure includes conditioning of the column
by flushing certain liquids through the column. The purpose of the conditioning is to make
the column ready for the application, and this will be discussed in more details below. The
second step is the sample loading, where the sample solution is flushed through the column.
During sample loading, the analytes are retained in the column by different interactions with
the stationary phase, whereas several of the matrix components are flushed out of the
column. In the next step, washing, one or more washing solutions are flushed through the
column. The purpose of the washing is to flush and remove as many of the matrix
components as possible, while the analytes are still retained in the column by strong
interactions with the stationary phase. This step gives sample clean-up, and is important to
eliminate potential interferences in the final chemical analysis. In the final step, elution, a
small volume of a certain liquid, which is called the eluent, is flushed through the column to
break the interactions between the analytes and the stationary phase, and to release the
analytes from the column. This final solution, which is called the eluate, is collected at the
bottom of the column. The eluate contains the analyte, and as few matrix compounds as
possible from the sample, and the eluate is used for the final chemical analysis.
18.7.2
The SPE Column
Figure 18.7 show a drawing of a SPE-column. The column itself is normally made of
polypropylene. A certain mass of stationary phase particles are packed into the bottom of the
column, and the stationary phase particles are held in place by two polyethylene filters on
each side of the bed. The particle size of the stationary phase particles is typically in the
range 40–50 mm. If you recall from the HPLC chapter (Chapter 13), the SPE particles are
much larger than the HPLC stationary phase particles, which are typically in the range
3–5 mm. The reason for using larger particles is that a simple vacuum at the column outlet is
used to obtain liquid flow through the system, whereas in HPLC a pump is used. Large
particles give a much lower back pressure, and it is easier to force the liquid through the
column. Because of the large stationary phase particles, one can easily flush biological fluids
with relatively high viscosity through the column. In addition, the high chromatographic
efficiency of small stationary phase particles is not required in SPE because the
“chromatographic separation” in SPE is very rough.
Sample Preparation
289
One can buy SPE columns with a lot of different dimensions depending on the
application, and loaded with different masses of stationary phase. The typical SPE column
for 1-ml plasma samples contains 30–100 mg of stationary phase. As discussed below, a
broad range of different stationary phases are available, but they are either based on silica
particles or organic polymeric particles. This means that the active part of the stationary
phase is functional groups, and these functional groups are attached to the surface of the
particles in large numbers. Silica-based particles have for many years
been very popular.
Those are highly porous particles with an average pore size of 60 A. Low molecular drugs
and other small molecules can diffuse into the pores and the surface in the pore structures
account for approximately 98–99% of the total surface of the particle. Thus, low molecular
compounds can be retained very efficiently. For macro-molecules in contrast, like proteins
in plasma and serum, their large size prevent them from entering the pore structures, and
they are just exposed to 1–2% of the particle surface. In other words, macro-molecules are
just slightly retained in the column, they are easily flushed from the column, and in this way
one can easily remove them from the sample by SPE.
18.7.3
Conditioning
As mentioned above, the first step in the SPE procedure is conditioning of the SPE column.
When SPE columns are obtained from a manufacturer, they are delivered in a dry state. In
the dry state, the functional groups are inactive and they have to be solvated before they can
retain our analytes. Solvation is carried out by flushing the column with a polar organic
solvent, and typically methanol is used. Figure 18.8 illustrates the effect of solvation. After
solvation, the functional groups have been “raised” and they are ready to retain the analytes
when introduced into the column. After flushing with methanol, the excess of methanol has
to be removed, which is located in the bed volume (¼ volume between the particles). The
reason for this is that methanol is a strong eluent in SPE, and if methanol is present in the
column when the sample is applied, the analytes can be poorly retained and partly lost from
the column. Methanol is normally removed in the second part of the conditioning process by
flushing with water or with aqueous buffers.
Conditioning is very important with silica-based SPE stationary phases, whereas for
some organic polymer-based phases, this is actually not required.
C18 functional groups
Silica particle
Conditioning
Silica particle
with methanol
Figure 18.8 Conditioning and solvation of a SPE stationary phase
290
Introduction to Pharmaceutical Chemical Analysis
Figure 18.9 SPE column and vacuum manifold
18.7.4
Equipment
SPE can be accomplished with relatively simple equipment as illustrated in Figure 18.9, and
the key elements are the SPE column and a vacuum manifold.
SPE columns are available from several manufactures. Normally, a SPE column is only
used for a single extraction, and then it is discarded. In addition, a vacuum manifold is
needed. The purpose of this is to make vacuum at the column outlet which force the liquids
through the column. At the connection between the bottom of the column and the
manifold, a valve is placed to turn on and off the vacuum within the column. The vacuum
is turned on when a liquid is to pass the column, whereas the vacuum is turned off in
between to avoid the column to dry out. In the vacuum manifold, small vials are placed
under each column to collect the final eluate from each extraction. In this way, several
samples can be extracted simultaneously.
18.7.5
Reversed-Phase SPE
Reversed-phase SPE is used for extraction of relative non-polar analytes from aqueous
sample solutions. Extraction is based on hydrophobic interactions between the analyte and
the stationary phase, and this is very similar to reversed-phase HPLC as discussed earlier in
this textbook (Chapter 13). This means that retention increases with increasing hydrophobicity of the analyte. The interactions are based on van der Waals forced between
Octadecyl C18
Si
C18H37
Octyl C8
Si
C8H17
Ethyl C2
Si
C2H5
Cyclohexyl
Si
Phenyl
Si
Cyanopropyl
Si
291
Increasing hydrophobicity
Sample Preparation
CH2CH2CH2CN
Figure 18.10 Different stationary phases for reversed-phase SPE
hydrogen–carbon bonds in the stationary phase and hydrogen–carbon bonds in the analyte
molecules. Hydrophobic interactions are promoted in aqueous or polar environment,
whereas they are broken in organic or nonpolar media.
Several different stationary phases are available for reversed-phase SPE as illustrated in
Figure 18.10, which summarizes some of the most popular ones based on silica particles.
The different stationary phases have different size of the hydrophobic group. C18 columns
are very popular, and they are also the most hydrophobic reversed-phase columns available.
Because of the high hydrophobicity, C18 columns give strong retention to many compounds.
This means that C18 columns are very general and can be used for many applications.
However, because it can extract a broad range of compounds, this column does not give the
best selectivity. In order to increase the selectivity, one can use some of the less hydrophobic
phases like C8 columns, which are also very popular.
In addition to silica-based SPE columns, one can also use polymer-based SPE columns
for reversed-phase SPE. One example of a polymer used as a stationary phase in SPE is
illustrated in Figure 18.11.
The stationary phase is a co-polymer build from two different monomers, namely the
hydrophobic monomer divinylbenzene and the more polar N-vinylpyrrolidone. The divinylbenzene moieties give the stationary phase hydrophobicity and these are responsible for
hydrophobic interactions with the analytes. In contrast, the N-vinylpyrolidone moieties give
the stationary phase polar functions in addition that can retain more polar compounds. An
important advantage of the polymer-based stationary phases is that they do not change
chemical properties even if they dry out during the procedure, and the initial conditioning
step can be avoided.
As mentioned above, hydrophobic interactions are promoted in aqueous or polar
environment, whereas they are broken in organic or nonpolar media. This means that a
relatively nonpolar drug present in an aqueous biological fluid is an ideal sample solution for
use with reversed-phase SPE. The aqueous sample can be loaded directly into the SPE
column. The nonpolar nature of the analyte results in strong retention by hydrophobic
292
Introduction to Pharmaceutical Chemical Analysis
N
O
Figure 18.11 Example of polymeric SPE stationary phase
interactions with the stationary phase. The aqueous environment promotes the hydrophobic
interactions between the analyte and the stationary phase. For acidic or basic analytes,
retention can be even better if the compounds are present in the sample in their uncharged
state. This means for basic analytes, alkaline conditions in the sample can improve the
retention of the analyte on the SPE column. This can be accomplished by diluting the sample
with an alkaline buffer.
During washing, pure water or other 100% aqueous solutions are normally used to
remove matrix components from the SPE column. This will remove water soluble
compounds having no affinity to the stationary phase, which means polar compounds.
By keeping the washing solution 100% aqueous, one still have very strong retention of the
analytes. In some cases, if the analytes are very well retained, aqueous washing solutions
containing 10 or 20% methanol or other organic solvents can be used, but care should be
taken not to start eluting the analytes from the column. Small amounts of methanol in the
washing solution remove more interference from the column and give a more clean extract.
In the final elution of the analytes, 100% methanol or mixtures of methanol and water are
typically used. Pure methanol gives a 100% organic medium in the column, and the
hydrophobic interactions are totally suppressed. The key point during elution is that the
hydrophobic interactions should be broken immediately in order to elute the analytes from
the column in the lowest volume of eluent as possible. The reason for using small volumes of
eluent is to avoid dilution of the eluate. If the analytes are basic or acidic compounds, pH in
the eluate can be adjusted to charge the analytes, as charging the analytes further reduce
hydrophobic interactions.
18.7.6
Secondary Interactions
With silica-based stationary phases, the functional groups are attached to the surface of the
silica particles. However, between the functional groups, residual silanols are present as
illustrated in Figure 18.12.
Sample Preparation
Secondary
interaction
293
C8 functional group
Hydrophobic interaction
Silica particle
H
N
O
H
Analyte
H
Residual silanol
Figure 18.12 Secondary interactions in SPE
Silanols are acidic, and in aqueous medium, they can retain analytes due to ionic
interactions. These interactions are called secondary interactions. Secondary interactions
can be advantageous, because analytes retained both by hydrophobic interactions and by
secondary interactions are strongly retained in the column. In such cases, stronger washing
solutions can be used without loosing the analyte, resulting in better sample clean-up. For
example, pure acetonitrile can be used during washing, and this solvent will remove many
both nonpolar and more polar matrix components, while the analyte is still retained by the
secondary interaction. To elute analytes retained by secondary interactions, methanol mixed
with an acid is normally used. Methanol suppresses the hydrophobic interactions, while the
acid reduces the ionization of the silanols and suppresses the secondary interaction,
facilitating the analyte to be eluted from the column.
18.7.7
Ion Exchange SPE
Analytes containing ionic groups, like acids and bases, can be extracted from aqueous
samples by ion exchange SPE. The principal type of interactions in ion exchange SPE is
ionic interactions between ionic groups on the stationary phase and ionic groups on the
analytes. Cationic analytes are positively charged and they are extracted with cation
exchangers, which are negatively charged. Many APIs are amines and they can be
extracted with cation exchangers. In a similar way, anionic analytes are negatively
charged, like carboxylic acids, and they can be extracted with anion exchangers,
which are positively charged.
Figure 18.13 gives an overview of some ion exchangers that are used frequently for SPE.
As illustrated, the stationary phases are characterized as strong ion exchangers or as weak
ion exchangers. A strong ion exchanger is ionized in the entire pH range, and will act as ion
exchanger regardless of pH in the SPE column. In contrast, a weak ion exchanger is only
ionized in parts of the pH range, and weak ion exchangers can be turned “on” and “off”
depending on pH in the SPE column (see also Section 11.7).
Strong cation exchangers typically use sulfonic acid-functionalities as the ionic group.
The sulfonic acids have very low pKa values, this means that they are strongly acidic and
therefore they are charged almost in the entire pH range. The strong cation exchangers work
as ion exchangers in the entire pH range. Weak cation exchangers are often based on
carboxylic acid-functionalities as the ionic group. Carboxylic acids are weak acids, with
294
Introduction to Pharmaceutical Chemical Analysis
-
Sulfonylpropyl
Si
CH2CH2CH2SO3
Strong cation exchanger
Carboxymethyl
Si
CH2COO-
Weak cation exchanger
(pKa = 4.8)
Trimethylaminopropyl
Si
+
CH2CH2CH2N —(CH3)3
Diethylaminopropyl
Si
+
CH2CH2CH2N (CH2CH3)2
Strong anion exchanger
Weak anion exchanger
(pKa = 10.7)
H
Figure 18.13 Overview of some stationary phases for ion exchange SPE
pKa values around 4.5–5.0. This means that, if pH in the SPE column is adjusted to above pH
3, the ion exchanger is gradually turned “on”, and if pH is reduced well below pH 3, the ion
exchanger is turned “off”. As discussed later, this can be utilized to control retention and
subsequent elution of analytes in ion exchange SPE.
Also anion exchangers can be divided into strong and weak materials. Strong anion
exchangers typically contain a quaternary ammonium group as the functional group, and
quaternary ammonium groups are ionized during the entire pH range. This means that strong
anion exchangers are always turned “on”. Weak anion exchangers normally contain a
secondary or tertiary amine as the functional group, and these have pKa values typically in
the vicinity of 10. Therefore, weak anion exchangers can be turned off by adjusting pH
above 10 in the SPE column, whereas it can be turned “on” if pH is lowered well below
pH 10. Figure 18.14 illustrate retention of amphetamine as analyte on a strong cation
exchange SPE column.
During sample loading in Figure 18.14, it is important that both the ion exchanger and the
analyte are charged to get ionic interactions and retention. Since a strong cation exchanger is
used in this case, it is charged (turned “on”) in the entire pH range. However, it is important
to be careful with pH in the sample, because during loading, amphetamine should be in its
charged form. Therefore, pH is adjusted well below the pKa value of amphetamine. The pKa
value of amphetamine is about 10.1, and it is generally recommended to adjust pH at least
Analyte
Ionic interaction
NH3
+
-
SO3
Silica particle
Sulfonylpropyl
Figure 18.14 Retention of amphetamine on a strong cation exchange SPE column
Sample Preparation
295
two units below the pKa value for basic analytes to be sure that the compound is totally
ionized. This means that pH in the sample related to Figure 18.14 should not exceed 8.0. If
pH adjustment is required in the sample, one prefers to use buffers because this gives
reasonable control of pH. In general, the same holds for acidic analytes. They also have to be
ionized during loading, and this requires a pH in the sample exceeding pKa þ 2.
In general, to ensure strong retention of the analyte during sample loading, it is also
important to have control on the content of other ions (matrix ions) in the sample. The reason
for this is that matrix ions at very high concentration, or with very high affinity for the ion
exchanger, can compete with the analyte for the active sites in the stationary phase. If this is
the case, the analyte can be partly flushed through the column during loading and can be lost.
This in turn will give low extraction recoveries. High content of matrix ions in the sample is
especially challenging when extracting urine samples, known to contain many different ions
which can vary in concentration from sample to sample. In such cases, the sample can be
diluted with water prior to extraction in order to reduce the concentration of matrix ions.
After loading, the SPE column is washed in order to remove as many matrix components
as possible. Normally water or preferably buffer solutions are used at this stage. One has to
be careful with pH in order not to turn “off” the ionization of the analyte during washing. If
the ionization of the analyte is partly turned “off” during washing, the analyte will be lost.
This in turn gives low extraction recoveries. One can also wash with methanol because
methanol does not disturb the ionic interactions between the analyte and the stationary
phase. Washing with methanol can be important to remove neutral matrix compounds with
low polarity, as they can be retained in the column by hydrophobic interactions to the
hydrocarbon chain between the SPE particle and the ionic head-group.
In the final step, the analyte is eluted from the SPE column. In the example from
Figure 18.14 with amphetamine, one can elute the analyte with an eluent with pH 12
which ensures that the charge on amphetamine is turned “off”, and the ionic interactions
are immediately broken. Addition of methanol to the eluent is probably an advantage as
amphetamine also can have some hydrophobic interaction with the hydrocarbon chains
of the ion exchanger. Methanol will also increase the solubility of amphetamine in the
eluent. In general, elution in ion exchange SPE is based on turning “off” the ionic
interactions. With strong ion exchangers, the ionic interactions are turned “off” by
turning “off” the charge on the analyte. With weak ion exchangers, on the other hand, one
can turn “off” the ionic interactions by turning “off” the ionization of the ion exchanger.
In all of the cases, turning “off” the ionic interactions involve selection of appropriate pH
conditions in the eluent.
18.7.8
Mixed-Mode SPE
In mixed-mode SPE, both hydrophobic interactions and ionic interactions are used for
retention of our analytes. This is illustrated in Figure 18.15, using amphetamine as
an example.
This silica-based sorbent contains C8 groups for hydrophobic interactions with the
analyte, and strong cation exchanger groups for ionic interactions with the analyte.
Polymer-based SPE stationary phases can also be purchased, and one example is illustrated
in Figure 18.16. In this case, sulfonate groups have been included in the polymer to give
ionic interactions with cations.
296
Introduction to Pharmaceutical Chemical Analysis
C8
Hydrophobic interaction
NH
+ 3
Silica particle
-
Analyte
SO3
Ionic interaction
Sulfonylpropyl
Figure 18.15 Retention of amphetamine on a mixed-mode SPE column
During sample loading, hydrophobic interactions are normally used as the primary source
of retention. This is because hydrophobic interactions are less affected by variations in the
sample matrix than ionic interactions. During washing in the next step, one first washes with
a buffer that turn “on” the ionic interactions between the analyte and the stationary phase. pH
in this solution should ionize both the stationary phase and the analyte. Washing with this
buffer will also remove many polar matrix components. One can then wash the column with
pure methanol because our analyte also is retained by ionic interactions, which are not
broken by methanol. By washing with methanol, neutral matrix components of low polarity
can be removed. This is the reason why mixed-mode SPE gives very clean extracts from
biological fluids.
For elution of the analytes in mixed-mode SPE, one needs to turn “off” both the
hydrophobic interactions and the ionic interactions. The hydrophobic interactions are
SO3-
N
O
Figure 18.16 Polymeric based stationary phase for mixed-mode SPE
Sample Preparation
297
suppressed by using methanol as one of the constituents in the eluent. The ionic interactions
are suppressed by adding either a base or an acid to the eluent. For basic analytes, base is
added to the eluent. High pH in the eluent will suppress the ionization of the analyte, and
typically methanol mixed with ammonia is used as eluent.
18.7.9
Normal-Phase SPE
Normal-phase SPE can be used to isolate polar analytes present in non-aqueous sample
solutions. Normal-phase SPE is much less in use in the pharmaceutical laboratory than
reversed-phase SPE, but it is discussed briefly in this section to complete the introduction to
SPE. In normal phase SPE, the analytes are retained in the column by polar interactions with
the stationary phase. The polar interactions are the same as discussed under normal-phase
HPLC (Chapter 11), namely hydrogen bonding, dipole/dipole interactions, induced dipole/
dipole interactions, and p–p interactions. For an analyte to give polar interactions, it should
contain one or more polar groups in its structure. Thus, compounds containing hydroxyl
groups, amines, carbonyl groups, aromatic rings, double bonds, or heteroatoms like oxygen,
nitrogen, sulfur, or phosphorous can be extracted by normal-phase SPE. Figure 18.17
illustrates some sorbents frequently used in normal-phase SPE.
The functional groups are hydroxyl groups, amino groups, or cyanopropyl groups. Even
pure silica particles, without any external groups attached to it, work as stationary phase for
normal-phase SPE. In the latter case, the silanols on the surface of the silica particles serves
as the functional groups.
Normal-phase SPE frequently relies on hydrogen bonding between the analyte and the
stationary phase. Hydrogen bonding is typical when hydrogen is bonded to electronegative
atoms like oxygen or nitrogen. Especially, strong hydrogen bonding occurs between
hydroxyl- and amino-groups. The polar interactions are turned “on” in a nonpolar
environment, whereas they are turned “off” in a polar environment. This means that the
sample should be dissolved in a nonpolar or low-polarity organic solvent during sample
loading. One example of such a non polar solvent is n-hexane. For elution of the analyte, one
has to change to a much more polar solvent that itself can hydrogen bond to the stationary
Silica
Si
OH
Diol
Si
CH2CH2CH2OCH2CH–CH2
OH OH
Cyanopropyl
Si
CH2CH2CH2CN
Aminopropyl
Si
CH2CH2CH2NH2
Figure 18.17 Overview of some stationary phases for normal-phase SPE
298
Introduction to Pharmaceutical Chemical Analysis
phase, and which can elute the analyte from the column. One example of a polar solvent that
can be used to suppress hydrogen bonding interactions is methanol. Methanol can thus be
used for elution of the analytes.
18.8 Summary
Sample preparation is important in pharmaceutical analysis. The intention of sample
preparation is to clean up the sample prior to the final chemical analysis, to avoid
interferences from matrix components in the sample, to avoid that matrix components
in the sample contaminate the analytical instruments and to make the sample compatible
with the analytical instrument. In addition, sample preparation also serves to enrich the
analyte to a concentration level, which is detectable by the final analytical method. Sample
preparation is important in connection with biological fluids, when analyzing drugs in
plasma, serum, whole blood, or urine samples. For sample preparation of biological fluids,
protein precipitation, liquid–liquid extraction, or solid-phase extraction can be used.
Sample preparation is also very important prior to analysis of APIs and certain excipients
in finished pharmaceutical products (quality control). For quality control, solid–liquid
extraction or liquid–liquid extraction can be used, depending upon whether the sample is a
solid or a liquid.
19
Analytical Chemical Characteristics
of Selected Drug Substances
This chapter selects a number of drug substances in order to demonstrate the significance of
chemical characteristics like solubility, log P and pKa values and the impact these basic
physicochemical functionalities have on analytical chemical work.
19.1 Amitriptyline and Mianserin
Amitriptyline and mianserin are both antidepressants and examples of basic hydrophobic
drugs. Monographs of the hydrochloride of both substances are found in Ph.Eur. and in USP.
Their structures are shown in Figure 19.1.
Amitriptyline and mianserin both have a large hydrophobic skeleton with a basic
tertiary amino group as the functional group. The high log P values indicate that the
substances are very fat soluble. Mianserin has a lower log P value because it has two
tertiary amino groups and amitriptyline has only one. As the free bases both compound are
practically insoluble in water.
The solubility of the corresponding hydrochlorides can be found in the monographs of the
Ph.Eur. and USP. Amitriptyline hydrochloride is freely soluble in water, alcohol and
methylene chloride. Mianserin hydrochloride is sparingly soluble in water, slightly soluble
in alcohol and soluble in methylene chloride.
Water solubility of the two basic drug substances is pH-dependent. They have their
highest solubility in an acidic environment, and the solubility is significantly reduced when
the ionization of the tertiary amino group is suppressed. Amitriptyline has a pKa value of
9.4. Mianserin has pKa values of 7.4 and 1.5. However, the latter value refers to an aromatic
tertiary amine and as such this group is a very weak base that is protonated only in strongly
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
Ó 2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
300
Introduction to Pharmaceutical Chemical Analysis
H3C
pKa = 7.4
N
pKa = 1.5
N
CH3
N
pKa = 9.4
CH3
Amitriptylin, log P = 4.9
Mianserin, log P = 3.8
Figure 19.1 Chemical structures of amitriptyline and minanserin with their log P and
pKa values
acidic environment. In aqueous solutions with pH > 12 both substances are unionized, and
the solubility is reduced at least 10 000 times relative to a neutral aqueous solution.
The salts of basic drug substances are more easily dissolved in water and other polar and
semipolar solvents like alcohols or acetonitrile. The information on solubility in the
pharmacopoeias therefore often includes solubility in water and in an alcohol like methanol
or ethanol. Nonpolar organic solvents are most often unable to dissolve salts, but there are
exceptions where hydrochlorides and similar salts of very hydrophobic molecules can be
dissolved in dichloromethane and amitriptyline and mianserin are such exceptions.
Ideal candidates for liquid-liquid extraction have high log P values. Liquid–liquid
extraction is therefore a good sample preparation method for amitriptyline and mianserin,
when extracted from an alkaline aqueous solution to an organic solvent. The distribution
ratio D between octanol/water at different pH values in water phase shows how the
extraction is affected by pH. Figure 19.2 shows the log D of mianserin as a function of
pH in the water phase.
4
.
.
.
.
.
3
log D
.
2
.
1
.
. .
−0.5
.
..
..
.
.
..
2
4
6
8
10
12
14
pH
Figure 19.2 Log D of mianserin as a function of pH
Analytical Chemical Characteristics of Selected Drug Substances
301
At pH > 9.4 the ionization of mianserin is less than 1% and log D becomes equivalent to
log P. Logarithm of the distribution ratio at this pH is 3.8, giving a distribution ratio of 6309.
Decreasing pH, so that the first tertiary amino group is ionized, reduces the distribution
ratio. At pH 7.4 this group is 50% ionized. A further decrease in pH also decreases the
distribution ratio because the ionized substance become more soluble in water. At pH 1.5
also the second tertiary amino group is ionized to 50% and log D is about 0, corresponding to
an equal distribution between the water phase and the organic phase.
In reversed phase HPLC, the substances have a strong retention, and as most silica based
column packing materials should not be used in the basic environment, the mobile phase is
typically acidic and the substances are chromatographed as cations. Since the drugs are very
hydrophobic, the mobile phase also has a relatively high content of organic modifier. The
tertiary amino group can provide ionic interactions with the silanol groups. In thin-layer
chromatography on silica, the mobile phase will typically contain an amine such as
diethylamine to suppress the ionic interactions that otherwise might induce peak-tailing
with reduced separation as the result.
Both substances absorb UV light. Amitriptyline has a maximum UV absorbance at
239 nm, where A (1%, 1 cm) is 504 and mianserin has maximum UV absorbance at 270 nm,
where A (1%, 1 cm) is 75. This low value is due to the chromophores being only isolated
benzene rings. In amitriptyline the chromophore of the benzene ring is extended with the
double bond in the side chain.
In Ph.Eur. amitriptyline is included as the hydrochloride, and the assay of this salt is
performed by titration with 0.1 M of sodium hydroxide using 96% ethanol as the solvent.
The ethanol is use as solvent to avoid precipitation of the base during titration.
Mianserin is also included in Ph.Eur. as a hydrochloride and also in this case the assay of
the drug substance is performed by titration with 0.1 M sodium hydroxide and using ethanol
as the solvent. Furthermore, 5.0 ml of 0.01 M hydrochloric acid is added to the sample
solution before titration and consumption of 0.1 M sodium hydroxide is read between the
two points of inflection.
19.2 Morphine and Codeine
Morphine and codeine are opium alkaloids. The formulas are shown in Figure 19.3 together
with the pKa values and log P values.
HO
H3CO
pKa = 8.2
O
N
N
CH3
HO
pKa = 8.2
O
CH3
HO
Morphine, log P = 0.9
Codeine, log P = 1.2
Figure 19.3 Chemical structures of morphine and codeine with their pKa and log P values
302
Introduction to Pharmaceutical Chemical Analysis
Both drugs are weak bases. The tertiary amino group has a pKa value of 8.2 for both
compounds. Both the tertiary amino group and the aliphatic hydroxyl groups contribute to
the substance’s polar character. Codeine has a log P of 1.2, while morphine has a log P of 0.9.
Log D values for the two compounds determined at pH 7.4 are 0.22 and 0.07, respectively.
Morphine has a lower log P value than codeine because it has a polar phenolic group where
codeine has a methoxy group. The phenolic group has a pKa value of 9.9. The low log P
value indicates that the substance is slightly soluble in water. Morphine has its lowest
solubility in water at pH about 9.1.
Standards for the hydrochlorides, phosphates and sulfates are given as monographs in
Ph.Eur. and USP. The salts are soluble in water and slightly soluble in alcohol. The solubility
of both drugs is pH-dependent. It is highest in the acidic environment where the amino
groups are ionized. The solubility of codeine is reduced when the ionization of the amino
group is suppressed. At pH > 11, ionization is completely suppressed and the solubility is
reduced approximately 1000 times compared to the acidic environment. The solubility of
morphine is also reduced when the ionization of the amino group is suppressed. But when
the pH becomes high enough the phenolic group will start to ionize, and thus increase the
water solubility again. At pH > 12, morphine is dissolved in water as the phenolate ion.
Both substances can be extracted from aqueous solutions using liquid–liquid extraction,
but one must take into account the pH of the water phase. Codeine is extracted from the
highly alkaline environment. Morphine is extracted best at ca. pH 9, which gives the least
ionization of both the amino group and of the phenolic group.
In reversed phase chromatography on silica-based columns a mobile phase with acidic
pH will often be preferred. In these systems the retention of morphine will be low and
therefore ion pairing can be used to increase retention. An anionic counter ion is added to
the mobile phase, and it will form ion pairs with the cationic analytes and in this way
increase the retention.
Both substances absorb UV light. In acidic aqueous solution, the maximum UV
absorbance for both morphine and codeine are at 285 nm. The substances also fluoresce
at 325 nm when being excited at 280 nm.
Codeine is included in Ph.Eur. as the free base, as the hydrochloride and as two
phosphates. The assay of the hydrochloride is performed similar to the procedure described
for mianserin hydrochloride (Section 19.1), while the assay of the free base and the
phosphates is performed by non-aqueous titration with 0.1 M perchloric acid.
Morphine is included in Ph.Eur. as the hydrochloride and as the sulfate. The assay of the
hydrochloride is performed similar to the procedure described for mianserin hydrochloride (Section 19.1), but the determination of the content of morphine in the sulfate is
performed by non-aqueous titration using 0.1 M perchloric acid and anhydrous acetic acid
as the solvent.
19.3 Ibuprofen and Naproxen
The examples of acidic drugs discussed are the anti-inflammatory drugs naproxen and
ibuprofen, the diuretic furosemide and the analgesic drug substance acetaminophen.
Ibuprofen and naproxen (Figure 19.4) are hydrophobic carboxylic acids. Monographs
of both substances can be found in pharmacopoeias. The high log P values indicate that the
Analytical Chemical Characteristics of Selected Drug Substances
303
CH3
O
CH3
pKa = 4.4
OH
H 3C
Ibuprofen, log P = 3.0
CH3
O
pKa = 4.4
OH
H3CO
Naproxen, log P = 3.7
Figure 19.4 Chemical structures of ibuprofen and naproxen with their log P and pKa values
substances have a very low solubility in water. In Ph.Eur. it is stated that they are
practically insoluble in water, soluble in alcohol and methanol, and that they are slightly
soluble in ethyl ether.
Water solubility is pH dependent. It increases with increasing ionization of the carboxylic
acid group. The pKa value of a carboxylic acid is typically in the range 4–5, and ibuprofen
and naproxen have pKa values of 4.4. Figure 19.5 shows how the solubility of naproxen
varies with pH. The solubility is about 0.03 g/l at pH 1 and about 188 g/l at pH 12. The
solubility increases approximately 6000 times from pH 1 to pH 12.
With a log P value of 3.7 ibuprofen is an ideal candidate for liquid–liquid extraction. If
ibuprofen is extracted from an acidic aqueous solution with an organic solvent, one can
expect the distribution ratio to be in the order of 1000. If the distribution ratio is 1000,
ibuprofen is 1000 times more soluble in the organic phase than in the water phase.
The high log P value indicates that ibuprofen is strongly retained in liquid chromatography at low pH on a reversed phase column. Maximum retention is obtained with an acidic
mobile phase with a pH that suppresses the ionization of the carboxylic acid group.
200
*
*
*
*
150
*
g/l
100
*
50
*
* * * * * * *
2
4
6
8
pH
10
12
14
Figure 19.5 Solubility of naproxen in water (g/l) as a function of pH
304
Introduction to Pharmaceutical Chemical Analysis
AU
225
250
275
300
325
350 nm
Figure 19.6 UV spectrum of ibuprofen in alkaline aqueous solution
The maximum UVabsorbance of both substances is in the range 260–280 nm. Figure 19.6
shows the UV spectrum of ibuprofen in alkaline aqueous solution. The maximum absorbance is at 264 and 273 nm. At 264 nm A (1%, 1 cm) is 18.5. The low value is originating
from UVabsorption of a single benzene ring. Naproxen has an A (1%, 1 cm) of 210 at 272 nm
and fluorescence at 353 nm when excited at 272 nm.
Ibuprofen is included in Ph.Eur. and the contant is determined by titration with 0.1 M
sodium hydroxide using methanol as the solvent. Naproxen is included in Ph.Eur. as such
but also as the sodium salt. The assay of naproxen itself is performed analog to the assay of
ibuprofen, while the assay of the sodium salt is performed using non-aqueous titration with
0.1 M perchloric acid and anhydrous acetic acid as the solvent.
19.4 Furosemide
The chemical structure of furosemide is shown in Figure 19.7.
Furosemide is a carboxylic acid with a pKa value of 3.9. The secondary aromatic amino
group is not protonated in the pH range 1–13. The sulfonamide group has a pKa value of
approximately 10. The log P value of furosemide is 3.0, which indicates that furosemide is
hydrophobic and that it is practically insoluble in water. It is soluble in alcohol and slightly
soluble in ethylether.
Water solubility is pH-dependent, and it increases when the carboxylic acid group is
ionized. In an aqueous solution at pH 1 the solubility is 0.01 g/l and at pH 10 the solubility is
45.0 g/l. Solubility has thus increased 4500 times in the alkaline environment.
Analytical Chemical Characteristics of Selected Drug Substances
pK a = 3.9
305
COOH
H
N
O
H2NO2S
Cl
Furosemide, log P = 3.0
Figure 19.7 Chemical structure of furosemide with log P and pKa values
The high log P value also indicates that furosemide has a strong retention in reversed
phase liquid chromatography. The highest retention is obtained in the acidic environment at
a pH which suppresses the ionization of the carboxylic acid group.
Furosemide may be extracted from an acidic aqueous solution by liquid–liquid extraction
using an organic solvent in which the substance is soluble.
UV spectra of furosemide dissolved in an acid and in an alkaline solution, respectively,
are shown in Figure 19.8.
The benzene ring system in furosemide contributes the most to the absorption of UV light.
It shows maximum absorption at two wavelengths at 274 and 342 nm in an acidic solution.
In an alkaline solution the maxima are shifted to be at 271 and 333 nm. In the acidic solution
A (1%, 1 cm) is 1333 at 274 nm.
In Ph.Eur. the content of the drug substance is determined using titration with 0.1 M
sodium hydroxide. Dimethylformamide is used as the solvent because furosemide is
practically insoluble in water and only sparingly soluble in ethanol. It is necessary to
carry out a blank titration to correct for possible acidic impurities in the solvent.
AU
Acid solution
Basic solution
225
250
275
300
325
350 nm
Figure 19.8 UV spectra of furosemide dissolved in an acid and in an alkaline solution
306
Introduction to Pharmaceutical Chemical Analysis
OH
pK a = 9.7
HN
CH3
O
Paracetamol, log P = 0.34
Figure 19.9 Chemical structure of acetaminophen with log P and pKa values
19.5 Paracetamol (Acetaminophen)
The chemical structure of acetaminophen is shown in Figure 19.9.
Acetaminophen is a phenol and has weakly acidic properties. The low log P value of 0.34
indicates that the drug substance is relatively polar, and that it will be somewhat soluble in
water. The Ph.Eur. monograph states that the substance is sparingly soluble in water, freely
soluble in alcohol and very slightly soluble in methylene chloride.
Paracetamol has two functional groups, a phenol group and an acetylated amino group.
The phenolic group can ionize, and the pKa value is 9.7. The N-acetyl-amino group is a
neutral amide group. This is typical of amide groups. In the pH range 1–8 paracetamol behave
as a neutral substance, and in this range the solubility is not affected by pH. The solubility in
water will increase with increasing ionization of the phenolic group, and the highest solubility
is obtained when the group is fully ionized at pH > 12. At pH 7 the solubility in water is 9 g/l
and at pH 12 it has increased more than 100 times to approximately 900 g/l.
The low log P value indicates that liquid–liquid extraction is not an efficient extraction of
the compound and a large excess of solvent must be used to get high yield of extraction. The
yield will not change in the pH range 1–8. Paracetamol cannot be extracted from alkaline
solutions because the phenolic group is ionized. Log D is 0.34 at pH 7 and decreases to 2.8
at pH 14.
The low log P value indicates that the drug has a polar characteristic and that it has a
relatively weak retention in reversed phase chromatography. In reversed phase chromatography a weak mobile phase with a low content of organic modifier must be used to provide
acetaminophen the sufficient retention. Retention will not be affected by pH in the range 1–7.
The UV absorption spectrum of paracetamol is due to the benzene ring with the
substituents. The UV absorbance is affected, however, dependent whether the phenolic
group is ionized or not. The UV spectrum obtained from an acidic aqueous solution will be
different from one obtained from an alkaline aqueous solution as shown in Figure 19.10. The
maximum absorbance in the acidic environment is at 245 nm where A (1%, 1 cm) is 642 and
in the alkaline environment at 257 nm with A (1%, 1 cm) of 710.
Analytical Chemical Characteristics of Selected Drug Substances
307
AU
Acidic solution
Alkaline solution
220
240
260
280
300
320 nm
Figure 19.10 UV spectra of paracetamol dissolved in an acidic and in an alkaline aqueous
solution
The assay of paracetamol is performed as a redox titration after hydrolysis of the amide
group. The drug substance is converted to the p-aminophenol by acid hydrolysis in sulfuric
acid. The p-aminophenol is then titrated with 0.1 M cerium sulfate [the p-aminophenol is
oxidized and Ce(IV) is reduced to Ce(III)].
19.6 Neutral Drugs
Neutral drug substances can have very different structure. The example discussed below
is hydrocortisone, which is a corticosteroid. The characteristic of neutral drug substances is that
they do not contain any ionizable functional groups like amino or carboxylic acid groups.
The structural formula of hydrocortisone is shown in Figure 19.11.
HOH2C
CH3
HO
O
OH
CH3
O
Figure 19.11 Chemical structure of hydrocortisone
308
Introduction to Pharmaceutical Chemical Analysis
Hydrocortisone has a highly hydrophobic steroid skeleton with some polar hydroxyl and
keto groups attached. The polarity of steroids increases with the number of polar groups and
hydrocortisone has three hydroxyl and two keto groups. The log P value is 1.4.
The solubility of hydrocortisone in water is about 0.3 g/l and it is freely soluble in ethanol.
It is a neutral substance and the solubility in water does not change with pH. The polar
hydroxyl group reduces the solubility in organic solvents that are immiscible with water. It is
slightly soluble in chloroform and practically insoluble in ethyl ether.
Hydrocortisone can be extracted from aqueous solutions by liquid–liquid extraction
using an organic solvent that can dissolve the substance. The log P value of 1.4 corresponds
to a partition coefficient between octanol and water of 25.
Hydrocortisone will be retained well in reversed phase liquid chromatography. Retention
is not affected by the pH of the mobile phase.
Hydrocortisone absorbs UV light and it is the a,b-unsaturated ketone that is the
responsible chromophore. The maximum UV absorbance obtained in ethanol as solvent
is at 240 nm and A (1%, 1 cm) is 435. This physicochemical characteristic is also used in
Ph.Eur. for assay of the content of hydrocortisone.
20
Quantification and Quality
of Analytical Data
This chapter focuses on the various aspects of quantitative analysis including calibration
methods, the principles of analytical validation and system suitability testing. Calibration
methods such as the external standard method, the internal standard method, standard
addition and normalization are discussed in detail. The performance characteristics used
in the validation process are defined along with typical methods by which it may be
measured. Test parameters usually employed in system suitability testing are discussed at
the end of this chapter.
20.1 Peak Height and Peak Area
A chromatogram of an analyte is shown in Figure 20.1. In the beginning no detector
response is recorded and the straight line at zero detector response is called the baseline.
When the substance enters the detector the detector response increases until a maximum
value is reached and then returns to zero. Ideally the peak appears to be a symmetrical
Gaussian curve.
The peak height is measured as the distance from the baseline to peak maximum, and the
peak area is the area measured between the peak and the baseline. If conditions are properly
controlled; both peak heights and peak areas vary linearly with the concentration of an
analyte in the solution injected. Both peak heights and peak areas can therefore be used as
parameters to determine the concentration of the analyte.
The measurement of peak heights can be made with high precision and yields accurate
results, provided variations in column conditions do not alter peak width. As peak height is
inversely related to peak width then all parameters that can affect the peak width must be
held constant. The parameters that affect peak height are the retention factor, the number of
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
310
Introduction to Pharmaceutical Chemical Analysis
peak area
peak height
baseline
Figure 20.1 Peak height and peak area
theoretical plates and the symmetry factor. When the retention factor increases peak
broadening occurs and the peak height decreases. A reduction in the number of theoretical
plates and an increase of the symmetry factor also decreases peak heights.
Peak area is independent of peak broadening effects. From this standpoint, therefore,
peak area is a more satisfactory analytical parameter than peak height. In modern
instruments peak area is measured with a high degree of accuracy and precision, and
the peak area is usually the recommended parameter for quantitative determinations.
Peak heights and peak areas are dependent on the properties of the analyte and on the
properties of the detector.To determine unknown concentrations calibration curves must be
prepared for each analyte. The most important calibration methods are:
.
External standard method;
Internal standard method;
. Standard addition;
. Normalization.
.
20.2 Calibration Methods
20.2.1
External Standard Method
An external standard is a certified reference substance of the compound to be determined or
a substance that has similar high quality. In the external standard method a series of standard
solutions containing known concentrations of the reference substance are prepared separately from the sample solutions. These solutions are called external standards solutions and
their concentrations approximate the concentration of the unknowns. The standard solutions
are analyzed separately from unknown samples under identical conditions. The peak areas
or the peak heights of the reference standards are read from the chromatograms and are
plotted as a function of concentration. The calibration curve should yield a straight line. An
example of a quantitative determination based on external standard is shown in Box 20.1.
Quantification and Quality of Analytical Data
311
Box 20.1 Quantitative determination based on the external standard
method
The concentration of a compound in a sample solution should be determined. It was
expected that the unknown concentration should be in the range of 100–250 ng/ml.
Standard solutions containing 100, 150, 200 and 250 ng/ml of the reference standard
were prepared and a fixed volume was accurately injected into the chromatographic
column. The peak areas of the reference standard read from the chromatograms are:
Concentration of standard (ng/ml)
Peak area
100
150
200
250
110 952
159 143
221 645
270 045
The concentrations are plotted against the peak areas yielding a straight line:
The regression line calculated by the method of least squares is:
y ¼ 1080x þ 1523
where y is the peak area and x is the concentration in ng/ml.
The sample solution was analyzed at the same conditions as the standard solutions
and the peak area of the analyte was 175 432. The unknown concentration x of the
analyte was:
175 432 ¼ 1080x þ 1523
x ¼ ð175 432 1523Þ=1080 ¼ 161:0
The concentration was 161 ng/ml.
312
Introduction to Pharmaceutical Chemical Analysis
In the test of purity (assay) of an active ingredient by an LC method there is no need to
carry out a calibration plot with a range of standard solutions since the method has already
been demonstrated to be linear. In this case it is preferable to prepare all standard and sample
solutions at approximately the same concentration. The method is called one point
calibration and it is a preferred method in pharmacopeias for the assignment of purity.
A concentration for the standard representing 100% purity is chosen. Following the
injection of the sample and the standard solution and measurements of their peak areas
in the chromatogram, the concentration of analyte in the sample solution may be determined
according the following equation:
AS =AR ¼ CS =CR
ð20:1Þ
where AS is the peak area of the analyte, AR is the peak area of the reference standard, CS is
the concentration of the analyte and CR is the concentration of the reference standard. The
concentration of the analyte in the sample solution CS is:
CS ¼ CR AS =AR
ð20:2Þ
The determination of purity by one point calibration is demonstrated in Box 20.2.
The external standard method is a preferred method when fixed volumes of standard
solutions of known concentrations of reference standards are accurately injected into the
column. This is the case when the sample and the standard solutions are prepared by
dissolving a known quantity of a compound in a fixed volume of solvent. In other cases
extensive sample preparation that involves extraction of the analyte, evaporation of solvent
and reconstitution of the sample in another solvent, is necessary. In these situations sample
losses are likely to occur and the exact concentrationof the solution to be injected becomes
unknown. In those cases the external standard method cannot be used because it cannot
compensate for sample losses.
Box 20.2 Determination of purity by one point calibration
A standard solution was prepared by dissolving 103.5 mg of a certified reference
standard into 500 ml of solvent in a volumetric flask. A fixed volume of 25 ml was
injected into a HPLC column and the peak area was 153 885. A sample solution was
prepared by dissolving 99.6 mg of the compound into 500 ml of the same solvent in
another volumetric flask. A fixed volume of 25 ml was injected into the column and the
peak area was 147 906.
The concentration of reference standard in the standard solution was:
103:5=500 mg=ml ¼ 0:21 mg=ml
The concentration of compound CS in the sample solution was:
CS ¼ 0:21 147 906=153 885 mg=ml ¼ 0:19 mg=ml
The amount of compound in the sample solution was: 0.19 500 mg ¼ 99.5 mg
The purity of the compound was: 99.5/99.6 100% ¼ 99.9%
Quantification and Quality of Analytical Data
313
Internal standard
Analyte
Figure 20.2 Chromatogram of analyte and internal standard
20.2.2
Internal Standard Method
The internal standard method compensates for analytical errors due to sample losses and
variable injection volumes. It is similar to the external standard method in that solutions of
the reference standard are compared with solutions of the sample. The key difference is that
a compound called an internal standard is added at a known concentration to standard
solutions and to sample solutions prior to any sample treatment. After sample processing
and injection the peaks of the analyte and the internal standard are shown in the
chromatogram as illustrated in Figure 20.2.
The analyte/internal standard peak area or peak height ratio is used in the calibration
rather than the absolute values such as in the external standard method. The internal
standard method can improve precision when the dominant sources of error are related to
sample preparation or injection.Such errors affect both the internal standard and the
analyte peak in the same way, and the peak area or peak height ratios becomes unaffected.
For this method to work well it is important to choose a suitable internal standard. The
internal standard should match the analyte closely, but not completely.Ideally an internal
standard should:
.
Be recovered to a similar extent as the analyte in the sample preparation procedure;
Be separated from all other substances in the sample;
. Have a detector response that is similar to the analyte;
. Have a concentration that gives a similar peak height or area as the analyte.
.
Consequently, the internal standard should have very similar physicochemical properties as
the analyte, but not so similar that it cannot be accurately determined. The ideal internal
standard in LC-MS or GC-MS is an isotopically labeled version of the analyte. A labeled
internal standard matches the analyte and is accurately determined at another mass. Labeled
internal standards are expensive and in most situations compound analogs are used as
internal standards.The internal standard method is a key method in bioanalysis and is
occasionally used in quality control of active pharmaceutical ingredients.
In the internal standard method a series of standard solutions containing known
concentrations of the reference standard are prepared in the actual matrix. Standard
solutions for determinations of drug substances in plasma are prepared in drug-free plasma
314
Introduction to Pharmaceutical Chemical Analysis
and standard solutions for determination of drug substances in urine are prepared in drugfree urine. The concentrations should approximate the concentrations of the samples. Then
the same amount of internal standard is added to all standard solutions and to all sample
solutions. All solutions are treated equally in the sample preparation procedure and are
subsequently injected into the column. The peak areas of the analyte and the internal
standard are read from the chromatograms and the analyte/internal standard peak area ratios
in each solution are calculated. The peak area ratios of standard solutions are plotted as a
function of concentration. The calibration curve should yield a straight line. The peak area
ratios of unknown samples are related to the concentration, which is found on the calibration
curve as shown in Box 20.3.
20.2.3
Standard Addition
Standard addition is a calibration method used to determine the concentration of an analyte
in a complex analytical matrix that contains substances that may interfere with the detector
response of the analyte. Because of interference, a calibration curve based on pure analyte
samples gives an incorrect determination. Adding known concentrations of a chemical
reference standard of the analyte to the unknown solution can solve the problem. This
calibration principle is called standard addition. An example of a quantitative determination
based on standard addition is shown in Box 20.4.
20.2.4
Normalization
Normalization is a technique used to provide a quantitative analysis of a mixture that is
separated by a chromatographic method. It can be applied to peak areas and it is assumed
that the detector response (detector signal per gram of analyte) is the same for all
components of the mixture. The quantitative results are obtained by expressing the area
of a given peak (A) as a percentage of the sum of all peak areas of the components (Ai) in
the mixture:
A
Að%Þ ¼ P 100
ð20:3Þ
Ai
An example is given in Box 20.5.
20.3 Validation
In pharmaceutical analysis the recommendations on analytical validation are based on the
work of the International Conference on Harmonization of Technical Requirements for
Registration of Pharmaceuticals for Human Use (ICH). The document Validation of
Analytical Procedures: Text and Methodology is the core document on analytical validation.
According to the document the validation process, which should demonstrate that an
analytical procedure is suitable for its intended purpose, consists of a series of performance
characteristics as shown in Figure 20.3.
Each of the performance characteristics is defined in the next section along with typical
methods by which it may be measured.
Quantification and Quality of Analytical Data
315
Box 20.3 Quantitative determination based on the internal standard
method
It was expected that the unknown concentrations were in the range of 100–250 ng/ml.
Standard solutions containing 100, 150, 200 and 250 ng/ml of the reference standard
were prepared and added an internal standard to a final concentration of 150 ng/ml.
After sample preparation a fixed volume was accurately injected into the chromatographic column. The peak areas of the reference standard and the internal
standard were read from the chromatograms and analyte/internal standard peak area
ratios were calculated:
Concentration of
standard (ng/ml)
100
150
200
250
Peak area of
the standard
Peak area of the
internal standard
Peak area ratio
110 952
159 143
221 645
270 045
164 302
163 997
164 204
164 289
0.675293
0.970402
1.349881
1.643719
The calibration curve is a plot of concentration versus peak area ratios:
The regression line y ¼ 0.0066x þ 0.0102 where x is the concentration in ng/ml and y
is the peak area ratio was used for calculation of analyte concentration in sample
solutions. Sample solutions were added internal standard to a final concentration of
150 ng/ml and were processed equally to the standard solutions. The sample solution
was analyzed at the same conditions as the standard solutions. In one of the sample
solutions the peak area of the analyte was 175 432 and the peak area of the internal
standard was 164 105. The peak area ratio 1.069023 was used to determine the
unknown concentration x of the analyte:
1:069023 ¼ 0:0066x þ 0:0102
x ¼ ð1:069023 0:0102Þ=0:0066 ¼ 160:4
The concentration was 160 ng/ml.
316
Introduction to Pharmaceutical Chemical Analysis
Box 20.4 Quantitative determination based on standard addition
The concentration of an unknown sample solution is expected to be ca. 1.0 ng/ml. To
determine the unknown concentration four standard addition solutions (SAS) were
prepared from the sample solution. Volumes of 5.00 ml were taken from the sample
solution and added to four 10-ml volumetric flasks. SAS 1 was prepared by diluting
5.00 ml of the sample solution to 10.00 ml. SAS 2, 3 and 4 were prepared by mixing
5.00 ml of the sample solution with respectively 0.25, 0.5 and 0.75 ml of a reference
solution followed by dilution to 10 ml. The reference solution contained 10 ng/ml of a
chemical reference standard of the analyte. The detector responses, which are the sum of
the unknown concentration (x) and the added concentrations, were:
Solution
Concentration of analyte in
the standard solution (ng/ml)
Detector response
SAS 1
SAS 2
SAS 3
SAS 4
x
x þ 0.25
x þ 0.50
x þ 0.75
99 151
149 987
200 132
249 889
The standard curve is a plot of concentration versus detector response:
The calibration curve calculated according to the method of least squares was:
y ¼ 201072x þ 99376
where y is the detector response and x is the concentration (ng/ml) in the standard
solution. The unknown concentration in SAS 1 is found at y ¼ 0:
x ¼ 99 376=201 072 ¼ 0:494
The concentration in SAS 1 was 0.494 ng/ml. As SAS 1 was prepared by diluting 5 ml of
the sample solution to 10 ml, the unknown concentration, c, of analyte in the sample
solution was:
c ¼ 0:494x 10=5 ¼ 0:99
The sample solution contained 0.99 ng/ml of analyte.
Quantification and Quality of Analytical Data
317
Box 20.5 Quantitative determination based on normalization
A mixture of 10 components was separated and the peak areas of each component were
recorded. The sum of peak areas of the 10 components was 12 533 495 and the peak area
of component A was 957 512. The percentage of component A in the mixture was:
A ð%Þ ¼
957512
100 ¼ 7:64
12533495
The mixture contained 7.64% of component A.
20.3.1
Analytical Procedure
The analytical procedure refers to the way of performing the analysis. It should describe in
detail the steps necessary to perform each analytical test. This may include the sample, the
reference standard and the reagents, use of apparatus, generation of calibration curve and
use of the formulae for calculation.
20.3.2
Accuracy
The accuracy of an analytical procedure is defined as the closeness of the test results
obtained by the procedure to the true value. In the case of the assay of a drug substance,
Accuracy
Precision
Specificity
Detection limit
Validation of
analytical procedures
Quantification limit
Linearity
Range
Robustness
Figure 20.3 Analytical characteristics used in validation of an analytical procedure
Introduction to Pharmaceutical Chemical Analysis
318
accuracy may be determined by application of the analytical procedure to an analyte of
known purity, for example, a certified reference substance or a substance with a similar high
quality. In the determination of a drug substance in a formulated product, accuracy may be
determined by application of the analytical procedure to synthetic mixtures of the
formulated product (drug-free) to which known amounts of the drug substance (e.g., a
reference substance) have been added.
Accuracy should be reported in either of two ways:
.
The percentage of recovery by the assay of the known added amount of analyte in the
sample;
. The difference between the mean and the accepted true value.
The accuracy of the analytical procedure should be established across its entire range. The
ICH document recommend that accuracy should be assessed using a minimum of nine
determinations over a minimum of three concentration levels covering the specified range.
This is three concentrations and three replicates of each concentration. Box 20.6 shows an
assessment of accuracy.
20.3.3
Precision
The precision of an analytical procedure expresses the closeness of agreement among
individual test results when the procedure is applied repeatedly to multiple samplings of a
Box 20.6 Assessment of accuracy
The range of this method was 100–1000 ng/ml. Three standard solutions were prepared,
containing the 100, 500 and 1000 ng/ml of the certified reference substance. The purity
of the reference standard was 100% and the concentrations of the standard solutions were
the accepted true value. The concentrations of the standard solutions were determined
three times. The results were:
100 ng/ml
500 ng/ml
1000 ng/ml
98.5
97.4
99.9
Mean ¼ 98.6
496
499
498
Mean ¼ 498
995
997
999
Mean ¼ 997
Accuracy reported as percentage of recovery:
100 ng=ml standard solution : 98:6=100 100% ¼ 98:6%
500 ng=ml standard solution : 498=500 100% ¼ 99:6%
1000 ng=ml standard solution : 997=1000 100% ¼ 99:7%
Quantification and Quality of Analytical Data
319
homogenous sample. Precision is usually expressed as the standard deviation (s) or the
relative standard deviation (% RSD) of the mean (m) of a series of measurements:
% RSD ¼ s=m 100%
ð20:4Þ
According to the ICH document, precision may be considered at three levels:
.
Repeatability;
Intermediate precision;
. Reproducibility.
.
The repeatability expresses the precision under the same operating conditions over a short
interval of time. Normally, the same analyst with the same equipment carries out the analysis
within one day within the same laboratory.
Intermediate precision expresses within-laboratories variations. Normally different
analysts with different equipment carry out the analysis on different days. The intermediate
precision is also called ruggedness.
Reproducibility expresses the precision of a procedure between different laboratories in a
collaborative study.
The precision of the analytical procedure is determined by assaying a sufficient number
of aliquots of a homogenous sample to be able to calculate statistically valid estimates of
standard deviation or relative standard deviation. The assays are independent analysis of
samples that have been carried through the complete analytical procedure from sample
preparation to final test result.
According to the ICH document repeatability should be assessed using a minimum of
nine determinations covering the specified range for the procedure. This is three concentrations and three replicates of each concentration. Repeatability can also be assessed using
a minimum of six determinations at 100% of the test concentration. Box 20.7 shows
assessment of repeatability using six determinations at 100% of the test concentration.
The acceptance criteria for precision depend very much on the type of analysis.
Pharmaceutical quality control precision of better than 1% RSD is easily achieved for
compound analysis, but the precision for biological samples is more like 20% at the
concentration limits and 15% at other concentration levels.
Box 20.7 Assessment of repeatability using six determinations at 100%
of the test concentration
The concentration of a sample solution containing 500 ng/ml of reference substance was
determined six times. The results of the assays were:
496; 499; 498; 495; 499; 497 and 496 ng=ml
Standard deviation : 1:57 ng=ml
Mean concentration : 497 ng=ml
Relative standard deviation : 1:57=497 100% ¼ 0:32%
320
20.3.4
Introduction to Pharmaceutical Chemical Analysis
Specificity
Specificity is the ability to assess unequivocally the analyte in the presence of compounds
that may be expected to be present. Typically these might include impurities, degradation
products and matrix components. Lack of specificity of an individual procedure may be
compensated by other supporting analytical procedures.
The ICH document states that when chromatographic procedures are used, representative
chromatograms should be presented to demonstrate the degree of selectivity. Peak purity
tests may be used to show that the analyte chromatographic peak is not attributed to more
than one component. Peak purity tests are based on spectra, such as UV/visible spectra,
recorded by diode array detectors. If all the individual spectra recorded during elution of a
peak are identical, the peak is considered pure.
Selectivity studies should also assess interferences that may be caused by matrix
components. In bioanalysis, interferences caused by biological fluids must be examined.
The absence of matrix interferences for quantitative methods should be demonstrated by the
analysis of drug-free samples from at least five independent sources of control matrix.
20.3.5
Detection Limit
The detection limit of an individual analytical procedure is the lowest amount of analyte in a
sample, which can be detected but not necessary quantified as an exact value. The test is a
limit test, which substantiate that the amount of an analyte is above or below a certain level.
The detection limit is usually expressed as the concentration of an analyte in the sample.
Several approaches for the determination of the detection limit are possible, depending on
whether the procedure is non-instrumental or instrumental. For non-instrumental methods
visual evaluation may be used. The detection limit is determined by establishing the
minimum level of an analyte that can reliably be detected by visual evaluation.
For instrumental methods, which exhibit baseline noise, the detection limit is based on
signal to noise ratios (S/N ratios). Determination of the S/N ratio is performed by measuring
signals from samples with known concentrations of analyte with those of blank samples
and establishing the minimum concentration at which the analyte can be reliably detected.
A S/N level between 3 : 1 or 2 : 1 is generally considered acceptable for estimating the
detection limit.
In chromatography the detection limit is the injected amount that results in a peak with a
peak height (H) at least two or three times as high as the baseline noise level (h). This is
illustrated in Figure 20.4, where H is the height of the peak and h is the range of the noise.
The detection limit is at a S/N level of 2: S/N ¼ H/h ¼ 2. Calculation of detection limit is
shown in the next section (see Box 20.8).
Other approaches described by the ICH documents depend on the determination of the
slope of the calibration curve and the standard deviation of the response.
S/N = H/h = 2
H
h
Figure 20.4 Limit of detection at S/N ¼ 2
Quantification and Quality of Analytical Data
321
Box 20.8 Calculation of detection limit and quantification limit
A standard solution containing a reference standard at a concentration of 16 ng/ml was
injected into a chromatographic system. The peak height (H) read from the chromatogram was 14 cm and the noise level (h) was 1 cm.
The detection limit at a signal to noise level of 2 corresponds to a peak with a peak
height (H) of 2h. Since h ¼ 1 cm, the detection limit corresponds to a standard solution
with a peak height H of 2 cm. The concentration of a standard solution with a peak height
of 2 cm is:
16 ng=ml ð2 cm=14 cmÞ ¼ 2 ng=ml
The detection limit of the analyte is 2 ng/ml.
The quantification limit is at signal to noise level of 10, corresponding to a peak with a
peak height (H) of 10 1 cm ¼ 10 cm.
The concentration of a standard solution with a peak height of 10 cm is:
16 ng=ml ð10 cm=14 cmÞ ¼ 11 ng=ml
The quantification limit of the analyte is 11 ng/ml.
20.3.6
Quantification Limit
The quantification limit of an individual analytical procedure is the lowest amount of an
analyte in a sample, which can be quantitatively determined with suitable accuracy and
precision. The quantification limit is a parameter of quantitative determinations of low
levels of compounds in sample matrices. It is used particularly for the determination of
impurities or degradation products of active ingredients and formulated products and in
bioanalysis for quantitative determinations of low levels of drug substances in biological
matrices. For instrumental methods that exhibit baseline noise the S/N ratio may be applied.
A typical S/N ratio is 10 : 1.
In chromatography the quantification limit is the injected amount that results in a peak
with a peak height (H) ten times as high as the baseline noise level (h). This is illustrated in
Figure 20.5.
An example on the calculation of the detection limit and the quantification limit is shown
in Box 20.8.
If the required precision of a method has been specified, the quantification limit of an
analyte can be determined by injecting a number of samples with decreasing amounts of
analyte. The calculated % RSD is plotted against the analyte concentration as shown in
Figure 20.6. The RSD increases when the concentration decreases and the quantification
limit can be read on the plot. Normally, the RSD criterion for defining the quantification
limit is set to 20%.
20.3.7
Linearity and Range
The linearity of an analytical procedure is its ability (within a given range) to obtain test
results, which are directly proportional to the concentration (amount) of analyte in the
sample as shown in Figure 20.7.
322
Introduction to Pharmaceutical Chemical Analysis
S/N = H/h = 10
H
h
Figure 20.5 Limit of quantification at S/N ¼ 10
Linearity should be established across the range of the analytical procedure. It should be
evaluated by inspection of a plot of signals as a function of analyte concentration or content.
The signal can be absorbance in a spectroscopic procedure and peak area, peak height or
peak area or peak height ratios in a chromatographic procedure. If there is a linear
relationship, test results should be evaluated by appropriate statistical methods, for example,
Precision (% RSD)
Required
precision
Concentration
Limit of quantification
Figure 20.6 Relationship between concentration and precision
Quantification and Quality of Analytical Data
323
Response
Linear range
Quantification limit
Detection limit
Concentration
Figure 20.7 Definition of linear range
by calculation of a regression line by the method of least squares. The square of the
correlation coefficient and the regression line should be reported.
The ICH recommends that, for establishing linearity a minimum of five concentrations
should normally be used. It is also recommended that the following minimum specified
ranges should be considered:
.
Assay of an active substance or a finished product: 80–120% of the test concentration;
For content uniformity: a minimum of 70–130% of the test concentration;
. For determination of an impurity: 50–120% of the acceptance criteria.
.
For bioanalysis, the range is normally established from the quantification limit and to
somewhat above the highest expected concentration found in the samples. Evaluation of
linearity in an assay of an active substance is shown in Box 20.9.
The range of an analytical procedure is the interval between the upper and lower
concentration (amount) of analyte in the sample (including these concentrations) for
which it has been demonstrated that the analytical procedure has a suitable precision,
accuracy and linearity.
The range of the procedure is validated by verifying that the analytical procedure provides
acceptable precision, accuracy and linearity when applied to samples containing analyte at
the extreme range as well as within the range.
20.3.8
Robustness
The robustness of an analytical procedure is a measure of its capacity to remain unaffected
by small, but deliberate variations in method parameters and provides an indication of its
reliability during normal usage.
324
Introduction to Pharmaceutical Chemical Analysis
Box 20.9 Evaluation of linearity in an assay of an active substance
Standard solutions of a certified reference substance were prepared across the
concentration range 80–120% of the test concentration. The standard solutions
were injected into a liquid chromatograph and the peak areas read from the chromatograms were:
Concentration of reference substance (%)
Peak area
80
90
100
110
120
80 102
90 037
10 0539
109 842
120 317
Linearity across the range of the analytical procedure is established by visual inspection
of a plot of peak area as a function of analyte concentration as shown below.
The regression line calculated by the method of least squares is:
y ¼ 1002x 67
where y is the peak area and x is the concentration. The square of the correlation
coefficient R2 ¼ 0.9997 show that these variables are highly correlated.
The evaluation of robustness should be considered during the development phase and
depends on the type of procedure under study. In the case of liquid chromatography,
examples of typical variations are:
.
.
Influence of variations of pH in the mobile phase;
Influence of variations in mobile phase composition;
Quantification and Quality of Analytical Data
325
.
Different columns (different lots/or suppliers);
Temperature;
. Flow rate.
.
In the case of gas chromatography, examples of typical variations are:
.
Different columns (different lots/or suppliers);
Temperature;
. Flow rate.
.
20.3.9
Test Methods in the European Pharmacopeia
The test methods given in monographs and general chapters of the Ph.Eur. have been
validated in accordance with accepted scientific practice and current recommendations on
analytical validation. Unless otherwise stated in the monograph or general chapter,
validation of the test methods by the analyst is not required.
20.4 System Suitability
One consequence of the evaluation of robustness should be that a series of system
suitability parameters is established to ensure that the validity of the procedure is
maintained whenever used. Typical variations are the stability of analytical solutions,
different equipment and different analysts. For chromatographic methods the variations
listed under robustness are crucial.
System suitability is an integrated part of many analytical procedures. The tests are based
on the concept that equipment; electronics, analytical operations and samples to be analyzed
constitute an integral system that can be evaluated as such. System suitability tests
parameters to be established for a particular procedure depend on the type of procedure
being validated. In Ph.Eur. system suitability tests are listed in the monographs.
For a chromatographic procedure the parameters that are usually employed in assessing
the performance of the columns are:
.
Number of theoretical plates;
Retention factor;
. Resolution;
. Relative retention;
. Symmetry factor.
.
In Ph.Eur. and USP the requirements for system suitability testing of HPLC methods are
typically:
.
The relative standard deviation of peak areas or peak heights are less than 1% (for drug
substance assay).
. The resolution (Rs) is greater than 2.
. The symmetry factor is in the range 0.8–1.5.
. The number of theoretical plates are larger than 2000 (for HPLC).
. The retention factor (k) is larger than 2.
326
20.4.1
Introduction to Pharmaceutical Chemical Analysis
Adjustment of Chromatographic Conditions
Chromatographic procedures in Ph.Eur. monographs contain necessary information that
enables the procedure to be reproducible in a laboratory and typically includes system
suitability requirements. In some instances, the chromatographic system fails to meet those
requirements. Because this situation can arise, adjustments to the operating conditions
might be necessary to meet the system suitability requirements. The maximum variations
allowed for each chromatographic parameter is found in the section on Chromatography.
Users are allowed to implement these adjustments to bring the system into suitable
performance without a full validation of the analytical procedure.
21
Chemical Analysis of Drug Substances
This chapter focuses on how raw materials, active pharmaceutical ingredients and
excipients for pharmaceutical production are controlled by chemical analysis to ensure
that they are of sufficiently high quality. The requirements of their identity and purity is
given in monographs in the pharmacopoeias (e.g., Ph.Eur., USP) where detailed tests to
be performed are described. We briefly review what pharmaceutical products are and how
they typically are produced. Then we discuss how the chemical quality control of raw
materials is performed, that is: (i) how to identify a test material to verify that it is the
correct ingredient for a given production, (ii) how to make sure that the raw material does
not contain significant impurities and (iii) how to determine the quantitative content of the
raw material.
21.1 What is a Pharmaceutical Raw Material, how is it Produced and why
must it be Controlled?
All active pharmaceutical ingredients (API) and excipients to be used in the manufacture of
pharmaceutical preparations are called pharmaceutical ingredients or pharmaceutical raw
materials. There are today a very large number of pharmaceutical raw materials, and the
most commonly used can be classified as given in Table 21.1. In the table examples from
each of the two groups of raw materials are shown and it is indicated whether the materials
are mainly used as an active ingredient, an excipient, or both.
This book primarily focuses on pharmaceutical raw materials that are well defined
chemical compounds or mixtures of organic compounds. The reasons for this are: (i) these
types of raw materials are the main ingredients in pharmaceutical preparations and (ii) the
use of chemical analytical methods constitutes a predominant part of their quality control.
The two groups can roughly be subdivided as shown in Table 21.2. For the other raw
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
328
Introduction to Pharmaceutical Chemical Analysis
Table 21.1 Overview of the pharmaceutical ingredients
Raw material
Example
API/excipient
Well defined chemical substances
Ibuprofen
Benzoic acid
Polyethylene glycol
Soya oil
Bentonite
Talc
Oxygen
Water
Technetium
Insulin
Coagulation factor
Human plasma
Herbal drugs
API
Excipient
Excipient
Excipient
Excipient
Excipient
API
Excipient
API
API
API
Excipient
API
Mixtures of organic substances
Mixtures of inorganic substances
Gasses and liquids
Radioactive substances
Biological macromolecules
Products of human or animal origin
Plant materials
materials given in Table 21.1 other techniques such as microscopy, physical testing,
biological tests and radioactivity measurements are used, and these techniques are outside
the focus of this book.
Raw materials that belong to the well defined group of chemical compounds are typically
organic chemicals produced by industrial synthesis on a relatively large scale. When going
from the starting material to the final product a number of synthetic steps are involved. An
example is shown in Figure 21.1 representing the active ingredient ibuprofen. When
performing synthesis of pharmaceutical raw materials, it is well known that there can
be side reactions and that these side reactions can give rise to undesirable byproducts in the
final raw materials. For ibuprofen a number of possible byproducts/impurities are shown in
Figure 21.1. The syntheses of APIs are often performed in organic solvents or the final raw
materials are often crystallized from an organic solvent, and the raw materials can therefore
also contain traces of organic solvents. To remove byproducts and traces of solvents the
Table 21.2 Overview of pharmaceutical raw materials of the type well defined chemicals or
mixtures of organic substances
Type of substance
Example
Well defined chemical compounds
Organic compounds
Acetaminophen
Methyl parahydroxybenzoate
Inorganic compounds
Sodium chloride
Sodium fluoride
Mixtures of organic substances
Polysaccharides
Cellulose
Synthetic polymers
Polytethylene glycols/macrogols
Fatty oils
Soya oil
Hydrocarbons
Vaseline
API/Excipient
API
Excipient
Excipient
API
Excipient
Excipient
Excipient
Excipient
Chemical Analysis of Drug Substances
CH 3
H 3C
CH 2
Isobutene
NaK
+
CH 3
H 3C
Benzene
Isobutylbenzene (I)
CH 3
I +
CH 3
Cl
O
H 3C
O
Cl
CH 3
CH 3
O
O
CH 3
AlCl 3
O
CH 3
NaOEt
H3 C
Ethyl chloroacetate
O
4'-Isobutylacetophenone
Acetyl
chloride
III
(III)
CH 3
CHO
CH 3
N
CH 3
H 3C
OH
H 3C
2-(4-Isobutylphenylpropioaldehyde
(IV)
CH 3
CH 3
H+
O
H3 C
CH 3
1. NaOH
2. HCl
IV
329
NH 2
CH 3
1. NaOH
2. HCl
O
H 3C
OH
CH 3
O
H 3C
Ibiprofen
CH 3
OH
O
Bu
Bu n
CH 3
OH
O
Bu
O
O
i
CH 3
i
OH
OH
Bu n
O
Figure 21.1 Synthesis scheme of ibuprofen (top frame) and possible byproducts in the process
(bottom frame)
synthesis products are normally purified before it can be used as a pharmaceutical raw
material. The purification process will usually remove most of the byproducts and solvents,
but residues can be left over. These residues may reduce the quality of the pharmaceutical
raw material and can cause adverse effects in patients when using the final drug. In addition,
330
Introduction to Pharmaceutical Chemical Analysis
HO CH2 CH2 O H
m
Figure 21.2 The general chemical structure of polyethylene glycol, m being between 5
and 1000
the pharmaceutical raw materials can also be contaminated with small amounts of
substances derived from:
.
Production equipment used during the synthesis and cleaning up;
Chemical degradation of the raw material as a result of storage;
. Packing used for storage of the raw material.
.
As discussed further in this chapter it is very important to control these impurities during
the pharmaceutical production so that the products that are produced are of a high
quality.
Raw materials that belong to the well defined group of compounds also include inorganic
compounds such as sodium chloride, calcium phosphate and sodium fluoride.
Raw materials belonging to the group of mixtures of organic compounds often have a
more diverse composition. Synthetic polymers are manufactured from pure chemicals, and
an example of this is polyethylene glycol (PEG) as shown in Figure 21.2. PEGs – also called
macrogols – are produced by reaction between water and ethylene oxide under pressure and
with the presence of a catalyst.
Fatty oils and polysaccharides can be produced by isolation from plant materials. An
example of this is soya oil. Soya bean oil is produced by extraction from the seeds of Glycine
max or G. soya using hydrocarbons. Hydrocarbons are produced from mineral oil.
Raw materials within the group mixtures of organic compounds may vary in composition
depending on how the raw material is produced. Remnants of solvents from the production,
impurities from packaging, impurities from the production and degradation products that
can affect quality in a negative direction can be present. For synthetic products it is starting
materials, byproducts from the synthesis and degradation products that have to be considered as impurities, while for raw materials isolated from plant materials various other
substances that are undesirable can be present (e.g., herbisides, insecticides, mycotoxins,
etc.). It is important to keep control over all these types of impurities, as well as of the overall
composition of the raw materials.
21.2 The Pharmacopoeias – the Basis for Control of Pharmaceutical Raw
Materials
As discussed briefly in Chapter 2, the European Pharmacopoeia (Ph.Eur.) contains both
requirements (standards) and test methods for quality control and identification of most
pharmaceutical raw materials that are not covered by patent protection. USP and the British
pharmacopoeia also include standards and test procedures for finished products. The major
part of Ph.Eur. consists of monographs for APIs and excipients used as pharmaceutical
ingredients. Besides the monographs for the raw materials a number of general chapters are
also included and these are listed in Table 21.3.
Chemical Analysis of Drug Substances
331
Table 21.3 Main contents of Ph.Eur. 7th edn (2011)
General chapters (vol. 1)
1. General notices
2. Methods of analysis
3. Materials for containers and containers
4. Reagents
5. General texts
. General texts on microbiology
. General texts on biological products
. Statistical analysis of results of biological assays and tests
. Residual solvent
. Alcoholimetric tables
. Assay of interferons
. Table of physical characteristics of radionucleides
. Pharmacopoeial harmonization
. Polymorphism
. Control of impurities in substances for pharmaceutical use
. Characters section in monographs
. Reference standards
. Gene transfer medicinal products for human use
. Functionally related characteristics of excipients
. Recommendations of methods for dosage forms testing
General monographs
. Monographs on dosage forms
. Monographs on vaccines for human use
. Monographs on vaccines for veterinary use
. Monographs on immunosera for human use
. Monographs on immunosera for veterinary use
. Monographs on radiopharmaceutical preparations and starting materials for
radiopharmaceutical preparations
. Monographs on suture for human use
. Monographs on herbal drugs and herbal drug preparations
. Monographs on homoeopathic preparations
Monographs (vol. 2)
Monographs
Index
Ph.Eur. begins with a general chapter that explains how the book and the monographs are
structured. In Chapter 2, a collection of all the technical methods used in the pharmacopoeia
is presented. In each method a short introduction to the principles of the methods is given.
Furthermore, the type of equipment to be used is described, including the demand on the
equipment and how to control that the equipment is functioning correctly. This will ensure
that the control of raw material is performed using proper equipment that does not give rise
to instrumental errors. The regulations and guidelines mentioned in Chapter 2 are the basis
of the chemical quality control of raw materials.
Chapter 3 provides monographs on materials to be used for manufacture of final packing
containers as well as on the packing containers themselves. The control of containers is
needed to ensure that components in the container do not leak into the medicine and that the
medicine is sufficiently protected against moisture and oxygen from the outside.
332
Introduction to Pharmaceutical Chemical Analysis
Chapter 4 defines all reagents and solutions used in the tests, and their preparation is
described in detail. This is very important in order to ensure that solutions are prepared
properly and in exactly the same way each time. Chapter 5 provides different types of
information. Besides statistical methods for processing data from analysis and requirements
regarding solvent residues in products a number of general texts are presented. These
texts contain information needed for the user of the pharmacopoeia and should be read
before performing tasks within the relevant area. For the chemical quality control the chapter
on “control of impurities in substances for pharmaceutical use” is of special importance.
Volume 2 of Ph.Eur. contains monographs (standards) for API and excipients with all the
tests needed to ensure high quality of the raw materials.
Each monograph for chemical compounds is structured to contain the same subsections:
.
Definition;
Production (optional);
. Characters;
. Identification;
. Test;
. Assay;
. Storage;
. Impurities.
.
An example of such a monograph is shown in Figure 21.3 for the active ingredient
paracetamol. Paracetamol is a well defined chemical compound, and the monographs of
this type of raw materials typically contain the following paragraphs:
At the top of each monograph we find the English trade name for the compound
(“Paracetamol”). In addition to this the Latin name (“Paracetamolum”) is also included.
After this we find a chemical structural formula (see Figure 21.3), the gross formula
(C8H9NO2 for paracetamol) and a molecular weight (Mr ¼ 151.2 for paracetamol).
The section on definition contains the requirements for actual content (from 99.0 to
101.0% for paracetamol). The limits for the content can also be considered to be the purity
requirement for the substance. The purity requirement of the compound is given in per cent
which is weight/weight%. Results obtained under assay have to be within these limits for the
API to be used for pharmaceutical production. That the upper limit can be above 100% is due
to the standard deviation on the analytical chemical method used. When titration methods
are used for the assay the limits are often 100 1%, while it is 100 2% when using
HPLC and 100 3% when using UV spectrophotometry. If a given compound is allowed to
contain some impurities in significant amounts it can influence the lower limit of the purity
requirement to be reduced.
If assay results obtained is outside the limits given it does not necessarily mean that there
is something wrong with the raw material, but it could equally mean that there is something
wrong with the performance of the quantitative analysis. In any case clarity has to be brought
to the issue before the raw material can be released to production.
The purity requirements do not refer to any trade name but rather to the systematic
chemical name to ensure a clear definition. For example, the raw material paracetamol
consists of 99.0–101.0% N-(4-hydroxyphenyl) acetamid which is the systematic chemical
name for paracetamol. Furthermore, the assay result has to be corrected for any content of
volatile impurities (loss on drying max. 0.5%).
Chemical Analysis of Drug Substances
333
Figure 21.3 Ph.Eur. monograph for paracetamol. Reproduced from The European Pharmacopoeia 7th Edition [2010] with permission of the EDQM
334
Introduction to Pharmaceutical Chemical Analysis
Figure 21.3 (Continued)
The section on characters always contains qualitative informations on appearance and
also most often on solubility. Besides these two characters the section can contain
information on polymorphism and in a few cases other physical characters.
The color and physical state of the substance is described under appearance. In addition
information on hygroscopicity can be given. If a substance is hygroscopic it has to be stored
in an air-tight container.
Chemical Analysis of Drug Substances
335
Table 21.4 Definition of solubility used under “Characters” in Ph.Eur
Description
Very soluble
Freely soluble
Soluble
Sparingle soluble
Slightly soluble
Very slightly soluble
Practically insoluble
Solubility (g/ml at 20 5 C)
>1.0
0.1–1.0
0.03–0.1
0.01–0.03
0.001–0.01
0.0001–0.001
<0.0001
Usually the solubility of the compound is given for avariety of solvents such as water, dilute
acid, dilute base, acetone and ethanol as well as in a nonpolar solvent, but also other organic
solvent can be used. Toxic solvents should be avoided. The solubility is discussed using
qualitative terms, as shown in Table 21.4. These terms are also used further in this book.
The information on solubility is given in the semi-quantitative terms “very soluble,”
“soluble” “slightly soluble,” and so on, which refer to some well defined intervals as shown
in Table 21.4.
The section on characters contains no requirements but rather qualitative information that
may be useful for a first hand impression of the physicochemical characteristic of the raw
material. Appearance and color give a quick indication of whether there is quality problems
related to a compound. Many pharmaceutical products are colorless compounds (white
powders), and if such products have a slight color, it may indicate that the raw material is
partially degraded. This means that the product has changed may be due to bad storage
conditions and thus the quality may be reduced. It is favorable to know the solubility
properties because the raw material normally has to be dissolved prior to analytical chemical
investigations.
The section on identification provides a description of how the raw materials should be
identified. It is of course of vital importance to verify that the content of a given container
holding a raw material is according to the labelling of the container. This is a key quality
control to ensure that the right ingredients are used for the production of a given drug.
Identity tests are not intended to provide a complete chemical identification (this would
typically require NMR spectroscopy), but rather a confirmation of the identity of the raw
material with an “acceptable” level of security. The technique of choice for identification is
IR spectrophotometry.
In some monographs the identification procedures are divided into “First identification”
and “Second identification”. The procedures under the “First identification” have to be used
in all circumstances. The test or tests in the “Second identification” may be used in
pharmacies and similar laboratories provided that it can be documented that the raw material
is fully traceable to a production batch that comply with all the other requirements of the
monograph.
After the section on identification a section for purity testing follows. These tests should
be performed to ensure that raw materials do not contain significant contamination of other
chemical compounds such as:
.
.
Starting materials from the chemical synthesis;
Byproducts from synthesis;
336
Introduction to Pharmaceutical Chemical Analysis
.
Contamination from industrial equipment;
Chemical degradation products;
. Contamination from packaging.
.
Most of the tests performed are relatively simple and does not intend to control for all
possible contaminants. The tests performed have been chosen to cover the major impurities
that characteristically can be found in the compound in question. This means that if other
impurities not covered by the tests are or can be present one must follow “good pharmaceutical practices” and not release the raw material to production before the “new”
impurities are examined. Section 21.5 reviews how to perform these tests.
The section for assay only contains one method for the quantitative determination
(“Assay”). This contains a procedure for how to quantitatively determine the content of
the raw material. In the “Assay” it is described both how the quantitative determination is
to be carried out experimentally, and how to calculate the purity of raw material. This
ensures that the quantitative determination is performed using a standardized procedure.
Results from the quantitative determination are compared with the requirement for content
under the “Definition.” If the result falls within the claim, and the raw material also fulfils
the tests under Identification and Tests, it has sufficient chemical quality to be used
for production.
The sections Tests and Assay complement each other and ensure that the raw material is
of high purity. The test for related substances is used for examination of the traces of related
organic chemical impurities at low concentration levels (typically 0.1% or less) which often
will not be reflected in the result of the assay. The quantitative determination, however, gives
a total and quantified overview of the purity of raw material. The quantitative method should
be highly accurate although it is not always specific but in combination with the tests for
purity it assures the quality of the raw material. Quantitative determination of raw materials
is reviewed in Section 21.6.
Some monographs contain a storage section at the end. In this section recommendations
for storage are given. If a compound is hydroscopic it is recommended to store it in an
airtight container. Light sensitive compounds should be stored protected from light.
Most monographs conclude with a paragraph that deals with “impurities.” This section
contains a list of systematic names and chemical structures of impurities typically found in
the raw material. The list is divided into two parts: the first part includes the specified
impurities which have a limit no more than what has been accepted by “the competent
authorities,” that is, what has been accepted at the time of registration or by later
certification. During the development of the drug a certain number and concentration
levels of impurities were present in the products tested in clinical trial. Such impurities are
therefore called “qualified impurities” as they have been tested to be of low toxicity in
clinical trials. These impurities are typically accepted in concentration levels of 0.1–0.5% in
the raw material. The second part of the impurities section covers “other detectable
impurities”. This is impurities with a known structure that can be detected in the given
test system. They are normally limited to less than 0.10%. These impurities are only given
for information. Impurities present in concentrations between 0.05 and 0.10% do not have to
be identified but should be included in the calculation of total amount of impurities.
Impurities being very toxic are treated on individual basis and limit can be as low as 0.1 ppm
for the impurity methylphenyltetrathydropyridine (MPTP) in pethidine.
Chemical Analysis of Drug Substances
337
The purity tests in the pharmacopoeias have until now been used as limit tests, but the test
for related substances will be changed into quantitative assays of the organic impurities in
accordance with the ICH guidelines Q3A, Q3B and Q3C. These guidelines have already
been adopted internationally and the pharmacopeia is now (2011) about to adopt the
principles.
This implies that a quantitative result of an impurity shall be given instead of a statement
like “passes test” as used with limit tests.
The above-mentioned requirements are for well defined chemical compounds. Monographs for mixtures of organic compounds like fatty oils and polymers are generally built up
in the same way. A significant difference is that such monographs often do not contain a
section on quantitative determination (“Assay”). However, it is customary to quantify one or
more components in the mixture. Methods for this are then given in the “Tests” section.
Furthermore, the monographs for mixtures of organic compounds do not contain chemical
structures, molecular formulas or molecular weight that is expected to be present for well
defined organic compounds. An example of a monograph for a mixture of organic
compounds is shown in Figure 21.4. In this case the castor oil is just defined as a fatty
oil, but under tests a gas chromatographic method is used to characterize the composition of
fatty acids in the product.
Following the section on characters the monographs contain a section on identification
(“identification”). The section provides a description of how the raw material is identified by
chemical analysis. In a number of cases the identity tests are not based on chemical analysis
methods but on a physical test such as measurement of viscosity. Identity tests are not
intended to provide a detailed chemical characterization of the raw material, but rather a
confirmation of the compound mixture to give an “acceptable” level of security of the
identification. The main identification procedures are reviewed in Section 21.4.
Monographs for ingredients that are mixtures of organic compounds, also contains a
section where various tests to determine whether there is contaminants or other types of
quality problems associated with the raw material (“Tests”). These tests can be based both
on chemical analysis and on other methods (biological or physical methods), and only the
main features of these are reviewed in Section 21.5. Monographs usually end with a
description of the recommendations relating to storage to prevent deterioration, and
recommendations regarding the labeling.
21.3 Which Contaminants are Found in Raw Materials, What are the
Requirements in a Maximum Content and Why?
Section 21.1 gives a brief discussion on the origin and nature of pharmaceutical raw
materials. The presence of impurities and contaminants is dependent on how the products
are produced and how they are stored. Section 21.2 gives an introduction to the standards for
pharmaceutical raw materials as they are described in pharmacopoeias and describes the
compilation of tests in the standards as given in the European Pharmacopoeia. The
philosophy of the pharmacopoeia is to control the quality of the raw materials used for
pharmaceutical drug preparation and in this way secure a basis for the quality of the final
drugs. If the raw materials are of high quality the final products are also more likely to be of
high quality depending on the production process.
338
Introduction to Pharmaceutical Chemical Analysis
Figure 21.4 Ph.Eur. monograph of Castor oil, virgin. Reproduced from The European Pharmacopoeia 7th Edition [2010] with permission of the EDQM
Chemical Analysis of Drug Substances
339
This Section discusses the types of impurities found in raw materials, the requirements
that generally apply to the presence of these impurities and why their presence is strictly
controlled.
21.3.1
Well Defined Chemical Compounds
Raw materials of the type well defined chemical compounds can typically be divided into
organic and inorganic compounds. Examples of a raw material of the organic type are
paracetamol and ibuprofen, while examples of raw materials of the inorganic type can be
sodium chloride and ferrous sulfate. The active pharmaceutical ingredient (API) in a
pharmaceutical preparation is, with few exceptions, an organic compound. Ferrous sulfate is
an example of an inorganic API. The group of excipients for pharmaceutical preparations
consist of both organic and inorganic compounds. Generally, the pharmacopoeias set up
very high demands for the purity of raw materials when it comes to the type well defined
chemical compounds, and requirements for content are typically within the range 97–103%,
and often even a smaller interval. The requirements for the content in nine different raw
materials are shown in Table 21.5 together with the method used in the quantitative assay.
The upper limit of the content reflects the ability of the assay method. The precision of a
method based on UV spectrophotometry is less than that of an HPLC method. Often
methods based on titration are preferred in the pharmacopoeias due to the high precision of
these methods. The titration methods are often less specific than the HPLC methods but a
high precision is preferred. The test for related substances is of course very important to
control the quality of the API, when an unspecific assay is used.
The lower limit of content reflect the precision of the assay method as well as the content
of accepted impurities and therefore it can be lower than expected when taking only the
precision of the assay method into consideration. In general the APIs are of a 98–99% purity,
leaving 1–2% for other contents.
This is illustrated in Figure 21.5, showing that 99.0–99.5% of the raw material shall
consist of the substance that is desired to be used in pharmaceutical production.
Even if the small fraction of impurities constitutes a very minor portion of the raw
material it is strictly controlled. The reason for this is that even minor amounts of impurities
Table 21.5 Examples of the purity requirements for content given in Ph.Eur. for a number of
APIs
API
Content (%)
Assay method
Omeprazole
Chlorpromazine hydrochloride
Simvastatin
Furosemide
Hydrocortisone
Ibuprofen
Paracetamol
Riboflavin
Budesonide
99.0–101.0
99.0–101.0
97.0–102.0
98.5–101.0
97.0–103.0
98.5–101.0
99.0–101.0
97.0–103.0
97.5–102.0
Titration
Titration
HPLC
Titration
UV spectrophotometry
Titration
Titration
UV spectrophotometry
HPLC
340
Introduction to Pharmaceutical Chemical Analysis
Figure 21.5 Visualization of the purity of a pharmaceutical raw material of the type well
defined chemical compound
can influence the quality and stability of the API as well as of the final drug. Table 21.6 shows
the major types of impurities found in organic and inorganic raw materials.
Earlier it was mentioned (Section 21.1) that organic compounds often contain impurities
having a similar structure as the API in question. These impurities are called related
substances and are mainly either starting materials, byproducts arising during the synthesis
or degradation products formed in the raw material during storage. The pharmacopoeias
require that the amount of such related substances should not exceed a certain percentage of
Table 21.6 Overview of common types of impurities in organic and inorganic raw materials of
the type well defined chemical compounds
API type
Impurity
Origin of impurity
Test to be performed
Organic
Related substances
(organic impurity)
Water
Synthesis/degradation
Related substances
Synthesis/absorption
from air
Synthesis
Production equipment
Synthesis/production
equipment
Synthesis
Water/loss on drying
Organic solvent
Heavy metals
Other inorganic
impurities
Minor ions
Inorganic
Inorganic
impurities
Heavy metals/
catalysts
Water
Synthesis/production
equipment
Synthesis/production
equipment
Synthesis/absorption
from air
Residual solvents
Heavy metals
Chlorides/sulfates/
nitrates/sulfated ash
pH/acidity or
alkalinity
Chlorides/sulfates/
nitrates
Heavy metals/specific
element test
Water/loss on drying
Chemical Analysis of Drug Substances
341
the raw material, and typically this number is 0.10% (w/w). In certain cases, when the
impurity is well known and it is nontoxic, the limit can be higher, but it can also be much
lower if the impurity is very toxic {e.g., the limit for MPTP in pethidine used for injection is
as low as 0.1 ppm [¼ 0.00001% (w/w)]}. The reason for the stringent requirements for
maximum presence of related substances is that:
.
They may have a pharmacological or toxic effect;
They can influence the effect of the preparation;
. They can influence the stability of the preparation.
.
How strict the requirements are depends on how potent the drug substance is. If the
maximum daily dose is larger than 2 g/day impurities above 0.05% (w/w) have to be
identified and qualified. Qualification implies that the toxicity of the impurity should be
clarified. For drugs given in less than 2 g/day the limit for identification is 0.10% (w/w) and
for qualification 0.15% (w/w).
In addition to related substances the pharmacopoeias set strict requirements for the
content of water in raw materials. Normally, the content of the compound provided under the
section “Definitions” is given based on the dried substance or anhydrous substance
dependent on whether a test for loss on drying or a test for water has been performed.
Thus the quantitative assay have to be corrected for the result obtained in the test for loss on
drying or the test for water before the result can be compared with the requirement for
content. In substances where no water should be present the limit is often 0.5% (w/w), but for
substances containing crystal water the limit will of course be accordingly and the limits will
be set taking into consideration how well defined the crystal water is. The water content has
to be controlled as it may:
.
Reduce the stability of the raw material during storage (there is more chemical reactions
taking place when there is moisture present);
. Result in a wrong dosing of the final drug if the water concentration is not taken into
consideration (a part of the compound consists of water and not 100% raw material).
In some cases, when the raw material is relatively unstable, the maximum limit for content
of water is more strict, and in other cases more than 0.5% (w/w) of water is allowed if the raw
material is very hygroscopic.
Traces of organic solvents remaining after production can be found in pharmaceutical raw
materials. These solvent is called residual solvents. Some organic solvent (e.g., benzene and
1,2-dichloroethane) are very toxic and should therefore not be present in APIs, while other
solvents like ethanol are less toxic. The solvents are divided into three different classes as
well as in a group of solvents for which no adequate toxicological data is known. Examples
of the solvent classes are given in Table 21.7. Solvent Class 1 is either carcinogenic or highly
toxic and should not be used in the production of pharmaceutical raw materials. Solvent
Class 2 is relatively toxic and there are strict requirements to the maximum content of these,
while the solvents in Class 3 are less toxic and thus are allowed in a little higher
concentration. Solvents in Class 2 are limited by their permitted daily exposure which
is tabulated, while the content of solvents from Class 3 solvents should be regulated through
GMP but should typically be below 0.5% (w/w). Limits for the individual organic solvent
are given in the pharmacopoeia.
Class 1 solvents (should be avoided)
Benzene
Class 2 solvents
Acetonitrile
N,N-dimethylformamide
Methanol
Class 3 solvents
Acetic acid
Ethylacetate
1,2-Dichloroethane
Cyclohexane
Ethyleneglycol
Tetrahydrofurane
1-Butanol
1-Propanol
Carbon tetrachloride
Chloroform
1,4-Dioxane
Pyridine
Acetone
Diethylether
Table 21.7 Examples of solvents in Classes 1, 2 and 3
Butylacetate
2-Propanol
Dichloromethane
Formamide
Toluene
1,1-Dichloroethene
Ethanol
N,N-Dimethylacetamide
Hexane
1,1,1-Trichloroethane
342
Introduction to Pharmaceutical Chemical Analysis
Chemical Analysis of Drug Substances
343
Table 21.8 Criteria for the inclusion of a heavy metals test in a monograph
Daily intake > 0.5 g/day, treatment < 30 days
Daily intake > 0.5 g/day, treatment > 30 days
Daily intake < 0.5 g/day, treatment > 30 days
Daily intake < 0.5 g/day, treatment < 30 days
Heavy metals test, limit 20 ppm
Heavy metals test, limit 10 ppm
Heavy metals test, limit 10 ppm if the
substance is used parenterally, otherwise
20 ppm
No heavy metals test
Heavy metals constitute another type of impurities that the pharmacopoeias by tradition
include a test for in the monographs. Previously this was an important test butnew productions
technologies and synthesis equipment of improved quality have resulted in this test to be
almost always negative. Ph.Eur. requires that the total amount of heavy metals present as
lead (Pb), copper (Cu), silver (Ag), mercury (Hg), cadmium (Cd), bismuth (Bi), rutenium
(Ru), gold (Au), platinum (Pt), palladium (Pd), vanadium (V), arsenic (As), antimony (Sb), tin
(Sn) and molybdenum (Mo) does not exceed the quantities shown in Table 21.8.
The test is based on the formation of the dark colored precipitate formed with
thioacetamide or sodium sulfide at about pH 3.5 using a known standard solution of
lead for comparison (see Section 21.5.1.8). However, this test is only valid for some of
the metals like Fe, Pb, Hg and Cu, while a number of the other elements are not detected.
Therefore this test is under discussion for revision into a method where also the modern
metal catalysts can be detected, and the use of modern instrumentation like ICP-AES or
ICP-MS is recommended.
Heavy metals can be found in raw materials as a result of contamination from industrial
equipment, or can be found in the raw materials that are derived from vegetable or animal
products (e.g., Hg in cod liver oil). The use of different catalysts in modern chemical
synthesis is a major challenge as these elements require a specific method. Heavy metals and
catalysts are hazardous and should be checked in raw materials where relevant. In addition, a
number of heavy metals can also initiate decomposition reactions.
The total content of inorganic material in well defined chemical compounds is normally
limited to 0.1% (w/w) of the raw material determined by the test for sulfated ash (see
Section 21.5.1.8). In addition some organic compounds must also be tested for the presence
of specific foreign inorganic ions such as chloride, sulfate or nitrate. This is done because
strong inorganic acids and bases are often used in the syntheses, and residues can be found in
the raw materials in spite of purification after synthesis. Tests for pH or acidity and alkalinity
are also used to verify protolytic residues from synthesis. The limits for the presence of
foreign ions vary by the same principle as given for heavy metals: The limits are stricter for
compounds to be given in relatively large daily doses, and they typically vary between 5 and
250 ppm (equivalent to 0.0005–0.025%).
The quality requirements for well defined inorganic compounds are essentially the same
as discussed above, and the limits are set by the same reasoning as for the well defined
organic compounds.
21.3.2
Mixtures of Organic Compounds
In substances considered to be mixtures of organic compounds a more diverse range of
contaminants are to be tested for. The reason for this is that this group consists of very
344
Introduction to Pharmaceutical Chemical Analysis
different ingredients, and that they can be manufactured in very different ways. Microcrystalline cellulose and carbomers are organic polymers, cetostearyl alcohol and hard fat are
waxes, polysorbate 80 and cetrimide are solubilizers and castor oil and cod liver oil are fatty
oils and together they are all examples of mixtures of organic substances. These mixtures
often have to be characterized using other test than those used for well defined organic
chemicals. The principle of what is allowed to be present in the raw material is basically the
same as discussed for well defined compounds. The higher the toxicity is, the more stringent
the requirements for the maximum content of a given impurity. The greater the expected
daily consumption is of the raw material in the finished product, the more stringent
requirements for maximum content.
21.4 How to Check the Identity of Pharmaceutical Raw Materials
21.4.1
Overview of the Identification Procedures
Previously, it was normal to have two sets of identification procedures for well defined raw
materials. The first set often included IR spectrophotometry for identification. For those
laboratories (e.g., pharmacies) not in possession of such an instrument a second set of
identification tests was needed. In new monographs, especially those on API not used in
pharmacies and hospitals, it is now becoming more usual only to have one test for
identification based on IR spectrophotometry. In Ph.Eur. identification procedures are
referred to as “First identification” and “Second identification” and an example is shown in
Figure 21.6 (mianserin hydrochloride) for the active ingredient. The procedures under “First
identification” can be used for a true positive identification while the procedures under
“Second identification” can be used, for example, in pharmacies provided that the substance
to be identified is fully traceable to a batch that comply with all the other requirements in the
monograph. The methods in the “First identification” is usually based on IR spectrophotometry which is a very specific method of identification. If the raw material supplied from a
vendor is documented to have been identified according to “First identification” it is only
necessary to perform “Second identification”. The “Second identification” procedures
usually require smaller and simpler equipment.
21.4.2
Techniques used for the Identification of Well Defined
Chemical Compounds
In connection with the identity testing of well defined chemical compounds, a large number
of analytical chemical techniques are used. The most important techniques are summarized
in Table 21.9. In addition a variety of color reactions and some other types of identity tests
are used. These tests are used only for one substance or for a very limited group of
substances, and are therefore not discussed further in this text.
21.4.2.1 Infrared Absorption Spectrophotometry
Infrared absorption spectrophotometry (IR spectrophotometry) is the most common
analytical technique used for identification of raw materials. The principle of IR
spectrophotometry is presented in Chapter 8. Using an infrared spectrophotometer, an
Chemical Analysis of Drug Substances
345
Figure 21.6 “First and second identification” for mianserin hydrochloride according to Ph.Eur.
Reproduced from The European Pharmacopoeia 7th Edition [2010] with permission of the
EDQM
Table 21.9 Overview of the main procedures used for the identification of well defined
chemical compounds
Analytical technique
Type of ID
Specificity
Infrared absorption spectrophotometry
Ultraviolet and visible absorption
spectrophotometry
Melting point
Polarimetry
HPLC
TLC
Color reactions
General identity reactions
First ID
Second ID
Highly specific
Relatively specific
First/second ID
Second ID
Second ID
Second ID
Second ID
Second ID
Low specificity
Low specificity
Relatively specific
Relatively specific
Low specificity
Low specificity
346
Introduction to Pharmaceutical Chemical Analysis
IR spectrum of the substance is recorded and the spectrum is compared with an IR
spectrum of a chemical reference substance (CRS) obtained on the same instrument. In
principle comparison could also be done using a reference spectrum recorded elsewhere,
but it can be difficult to compare IR spectra obtained on different IR instrument. The IR
spectrum can be obtained directly on the substance in solid, liquid or in gas phase.
Furthermore, it is sometimes preferred to record the IR spectrum from a solution. Solids
are often prepared by mixing with potassium bromide and pressing the mixture to a
transparent tablet. Potassium bromide is transparent to IR radiation in the wavelength
range used. An IR spectrum of a solid can also be recorded directly on the powder using
attenuated total reflectance technique. Liquids, if not too volatile, can be placed between
two sodium chloride plates. IR spectra of gasses can be obtained using a gas cell which of
course should have a longer path length to overcome the problem having fewer molecules
per unit length in the gas phase. A number of drug substances can show polymorphism. In
this case a compound can have more than one IR spectrum due to different crystal forms. In
these cases it can be necessary to recrystalize the substance in a suitable solvent, and the
CRS has to be treated similarly. If it is difficult to recrystalize the substance and the CRS
the possibility to obtain the IR spectra in solution should be considered. Identification of
nimodipine is an example of this (Figure 21.7).
In solution the crystal structure is destroyed and the spectra should be identical if the
substance and the CRS is the same compound. For the raw material to be identified
unambiguously the spectra have to be identical, which means that all the absorption band
in the two spectra are equal with respect to both the wavelengths (wave numbers),
absorption maxima (transmission minima) and with respect to relative absorption
intensities (transmittance). An example of the identification of a compound by IR
spectrophotometry according to Ph.Eur. is shown in Box 21.1. Spironolactone shows
polymorphism and if the spectra of the substance and the CRS are different a recrystallization in methanol is performed. The IR spectra in Box 21.1 show the IR spectra before
and after recrystallization.
For this drug IR spectroscopy is used for the “First identification.” The “Second
identification” is a TLC test and a color reaction.
IR spectrophotometry is a reliable method of identification. The probability that two
different compounds have identical IR spectra is very small, because the IR spectra
contain a large number of structure-specific absorption bands. IR spectra are like
fingerprints in the identification of chemical compounds. Due to its high security in
identification IR spectrophotometry is the preferred technique for “First identification”
for the majority of the monographs on well defined pharmaceutical ingredients. The
technique is simple to perform and it takes only few minutes to record the spectrum. It is
important when comparing IR spectra that they are recorded under virtually identical
conditions:
.
.
Recorded using the same sample preparation;
Recorded using the same infrared spectrophotometer.
It is also important to ensure that the IR spectrophotometer is working properly, which
means that the wavelength scale is calibrated in relation to a polystyrene film as discussed in
Section 8.4.
Chemical Analysis of Drug Substances
347
Figure 21.7 Identification of nimodipine using IR spectrophotometry according to Ph.Eur.
Reproduced from The European Pharmacopoeia 7th Edition [2010] with permission of the
EDQM
21.4.2.2 Ultraviolet and Visible Absorption Spectrophotometry
In addition to IR spectrophotometry, ultraviolet and visible absorption spectrophotometry
(UV spectrophotometry) can be used for idenfication of substances that absorb UV
radiation. The UV spectrophotometry technique is presented in Chapter 7. Using a UV
spectrophotometer a UV spectrum is recorded from a solution of the drug substance, and
absorption maxima and minima as well as absorption intensities are used for the characterization of the substance. To have a positive identification of the compound the UV spectrum
obtained must be in accordance with Ph.Eur. requirements both in terms of wavelengths at
the absorption maxima and with respect to absorption intensity. An example of identification of a compound by UV spectrophotometry is shown in Figure 21.6 for mianserin
348
Introduction to Pharmaceutical Chemical Analysis
Box 21.1 Identification of spironolactone according to Ph.Eur. 7th edn
(2011) by IR spectrophotometry
Infrared absorption spectrophotometry (2.2.24)
Comparison: spironolactone CRS.
If the spectra obtained in the solid state shows differences, dissolve the substance to be
examined and the reference substance separately in the minimum volume of methanol R,
evaporate to dryness and record new spectra using the residues.
Discussion of the procedure:
1–2 mg of spironolactone is taken out and mixed in a mortar with 300–400 mg KBr or
KCl. The two substances are mixed well and crushed thoroughly after which the powder
mixture is transferred to a tablet press and compressed into a tablet under vacuum. The
tablet is placed in an IR spectrophotometer and the IR spectrum is recorded where the
transmittance of IR radiation is plotted as a function of wave number. The IR spectrum of
the test sample is compared with the IR spectrum obtained of the CRS of spironolactone.
Other solid state techniques for obtaining IR spectra can be applied, for example,
attenuated total reflectance where the IR spectrum is recorded directly on the substance
without any sample preparation.
Requirements for positive identification:
The criterion for positive identification of spironolactone is that the IR spectra of
spironolactone raw material and of spironolactone chemical reference substance are
identical at least after recrystallization. IR spectra are shown in Figure 21.8.
hydrochloride. For this drug UV spectrophotometry is used for the “Second identification”
in the Ph.Eur. together with a TLC test and a test for chloride in Box 21.2.
UV spectrophotometry is a relatively easy technique to carry out in the laboratory, but
compared to IR spectrophotometry, it provides a somewhat lower trustworthy certainty in the
identification of a compound. This is because the UV spectra are considerably less detailed in
their structure and therefore less informative than the IR spectra, and a number of compounds
may have very similar UV spectra. For this reason UV spectrophotometry is only used in the
“Second identification” when used for identification. In Ph.Eur. it is common to determine
the wavelength of one or two absorption maxima or minima and to calculate the specific
absorbance at the absorption maximum as shown in Box 21.2. Sometimes also a ratio
between two maxima or a maximum and a minimum is calculated to give a characteristic
value for the compound. The rest of the UV spectrum is not used which is in contrast to the IR
spectra where identification is based on all details in the entire spectrum.
UV spectra are recorded on samples dissolved in a suitable solvent. It is important to use a
solvent that does not absorb UV light in the wavelength range in question. Water is very
suitable and non-UV-absorbing buffers (e.g., phosphate or borate) may be added to control
pH. But it is important that the UV spectrum is recorded under the conditions mentioned in
Figure 21.8 An IR spectrum of a sample of spironolactone are recorded (top left) and compared to spironolactone CRS (top right). Because the two
spectra are different the sample and the CRS are recrystallized in methanol and the IR spectra are recorded again (bottom). The IR spectrum of the
sample (bottom left) is now identical with the IR spectrum of the CRS (bottom right)
Chemical Analysis of Drug Substances
349
350
Introduction to Pharmaceutical Chemical Analysis
Box 21.2 Identification of mianserin hydrochloride according to Ph.Eur.
7th edn (2011) by UV spectrophotometry (see also Figure 21.6)
Discussion of the procedure:
50.0 mg of raw material is dissolved in water R and diluted to 50.0 ml with water in a
volumetric flask. 5.0 ml of this solution is diluted with water R to 50.0 ml in a new
volumetric flask. Then the UV spectrum is recorded in the range 230–350 nm.
Requirements for positive identification:
An absorption maximum should be observed at 279 nm and the specific absorbance A
(1 cm, 1%) at this wavelength should be in the range of 64–72.
Calculation Example:
If the absorbance (A) is measured to be 0.655 at 279 nm with a light path (b) of 1 cm the
specific absorbance can be calculated in the following way:
The concentration of mianserin hydrochloride after the first dilution (50.0 mg dissolved
in 50.0 ml):
C1 ¼ 50:0 mg=50:00 ml ¼ 1:0 mg=ml ¼ 0:10% ðw=vÞ
The concentration of mianserin hydrochloride in the second dilution used for the
measurement (5.00 ml diluted to 50.0 ml):
C2 ¼ 0:10% ðw=vÞ ð5:00 ml=50:0 mlÞ ¼ 0:010% ðw=vÞ
Using Beer’s law the specific absorption is calculated to be:
a ¼ A=b c ¼ 0:655=1 cm 0:010% ðw=wÞ ¼ 65:5
This value is within the requirement given for mianserin hydrochloride in Ph.Eur for a
positive identification (see Figure 21.6).
Comments on the procedure:
The procedure for how the test should be performed is thus described in the monograph.
The R behind water refers to the quality of water to be used. The quality requirements to
reagents including water R is described in a special section in Vol. 1 of Ph.Eur. The
general description of ultraviolet (UV) and visible (Vis) absorption spectrophotometry is
in Ph.Eur. given in Section 2.2.25.
The reason for weighing 50.0 mg raw material for the test is to reduce the uncertainty
of weighing as this uncertainty increases with decreasing amount. Stating 50.0 mg and
not just 50 mg is because this in principle is a quantitative test when the specific
absorption is determined, and therefore a higher accuracy is needed. The sample must be
diluted to give an absorbance in the range 0.2 to 0.8, and the dilution is done in two steps
to avoid the use of too large measuring flasks. In this case the concentration of the final
test solution is relatively high (10 mg/100 ml) which is necessary as the two isolated
benzene rings are the only UV-absorbing chromophores in the molecule.
Chemical Analysis of Drug Substances
351
the monograph, because the UV spectrum with its absorption maxima and its values for the
specific absorbance varies with the solvent, pH, ion strength and several other parameters as
discussed in Chapter 7.
If the substance is dissolved in a different way than prescribed, the requirements of
absorption maxima and specific absorbance will no longer be valid. Before performing the
identification using UV spectrophotometry, it is important that both the absorbance scale
and the wavelength scale of the UV spectrophotometer are checked and calibrated as
discussed in Section 7.7.
21.4.2.3 Thin-Layer Chromatography
Thin-layer chromatography is a third technique used for identification that often recurs in
the monographs of the Ph.Eur. Thin-layer chromatography is presented in Chapter 12 and an
example of a TLC test used for identification of mianserin hydrochloride is shown in
Figure 21.6. Similar to UV spectrophotometry TLC is also used for “Second identification”
(Box 21.3).
Box 21.3 Identification of mianserin hydrochloride by TLC according to
Ph.Eur. 7th edn (2011); see also Figure 21.6
Discussion of the procedure:
A test solution prepared by dissolving 10 mg of raw material in 5 ml of methylene
chloride (CH2Cl2). A reference solution (a) is prepared by dissolving 10 mg mianserin
hydrochloride CRS in 5 ml of methylene chloride. A reference solution (b) is prepared by
dissolving 10 mg of mianserin hydrochloride CRS and 10 mg of cyproheptadine
hydrochloride CRS in 5 ml of methylene chloride. 2 ml of each solution is applied to
the TLC plate provided with a coating of silica gel as the stationary phase. The elution is
performed using a mixture of diethylamine, ether and cyclohexane in the volume ratio
5 : 20 : 75. After development to two-thirds of the plate length the TLC plate is removed
from the development chamber and dried. For detection the TLC plate is viewed in
ultraviolet light at 245 nm. The reference solution (b) is used for system suitability and
two clearly separated spots should be visible.
Requirements for positive identification:
The main spot of the raw material should have the same position and size compared to the
spot in reference solution (a).
Comments on the procedure:
Raw material and reference material is dissolved in methylene chloride because this is a
good solvent for mianserin hydrochloride, and because it is easy to evaporate when
applying the solutions to the TLC plate. Silica gel is used as the stationary phase because
silica provides appropriate retention of mianserin via hydrogen bonding interactions
with the amino group. Diethylamine is added to the eluting solvent to reduce the ionic
interaction of the analyte with charged silanol groups on the stationary phase.
352
Introduction to Pharmaceutical Chemical Analysis
A number of different compounds may in TLC have similar affinity to the stationary
phase, and will therefore obtain about the same Rf value on the TLC plate. In spite of this
TLC provides considerable specificity because, in addition to the Rf value it is also possible
to compare the color of spots after derivatization as well as the intensity of the spots. The
technique is not used as the only identification test, but must be combined with other
methods. TLC is therefore most often only used in the “Second identification” series. The
method is relatively quick to perform and is cost effective because it requires only little
equipment and reagents.
21.4.2.4 Melting Point
Melting point is an identification technique that is used in some monographs. The principle
of this relatively simple technique is described below.
A small amount of raw material to be tested is inserted in a small capillary tube of glass
closed in one end by melting in a flame. The capillary with the sample is placed in an oil bath.
The temperature of the oil bath is increased typically by 10 C/min to 10 C below the
melting point and then the temperature is increased carefully at 1 C/min until the last of the
raw material has melted. The melting of the raw material is observed closely through a
magnifying glass, and the melting point can be read as the temperature where the last
remnant of the raw material has melted. The melting point can also be determined
automatically by instrumental techniques. A schematic diagram of a melting point
apparatus is shown in Figure 21.9.
Mixing
device
Thermometer
Capillary
tupe
Sample
Oil bath
Heater
Figure 21.9 Diagram for an apparatus for melting point determination
Chemical Analysis of Drug Substances
353
Box 21.4 Identification of paracetamol according to Ph.Eur. 7th edn
(2011) by melting point determination (see also Figure 21.3)
Melting point (2.2.14): 168 C to 172 C.
Discussion of the procedure and requirements for positive identification:
The melting point should be determined and must lie within the range 168–172 C.
Melting point determination has only a limited value as an identity test because many
different compounds can have melting points in the same range. Therefore, the method
cannot be used alone as an identification test, but must be combined with other methods
and is often used during the “Second identification”. The method is quick and inexpensive
to perform. It is important that the test is carried out on dried raw material. If water is
present in the raw material it can affect the melting point determination significantly.
Ph.Eur. Section (2.2.14) gives a description of the requirements to the apparatus used for
melting point. This is not discussed in detail here, but the requirements stated are
important to ensure that the melting point determination is carried out correctly.
When using determination of melting point for identification the result obtained is
compared to the requirement given in the monograph in question. The melting point is
determined as the temperature where the last remnant of the raw material has melted. If the
melting point falls within the claim the identity test is positive. An example of such an
identity test of paracetamol is shown in Box 21.4 (see also Figure 21.3). For this drug
substance the melting point determination is used for identification in the “First
identification” series (together with IR spectrophotometry) as well as in the “Second
identification” series (together with UV spectrophotometry, a color reaction and a test for
acetate).
21.4.2.5 Polarimetry
Optical rotation of the plane-polarized light is used to characterize compounds with one or
more chiral centers. The test may also be used as one of the tests for identification of such
compounds.
A number of pharmaceutical raw materials are chiral compounds, which means that they
contain one or more carbon atoms bound to four different atoms or groups as shown below.
Stereochemistry is also discussed in Section 3.3.
A carbon atom bound to four different groups (ligands) is called a chiral center, and the
spacious configuration of the four ligands make two isomers possible being mirror images of
one another similar to the right and left hands being mirror images of each other
(Figure 21.10).
The two enantiomers of the same material have identical physical properties as melting
point, boiling point, density and color (i.e., they have the same physicochemical properties),
but they can have very different pharmacological effects. A special feature of enantiomeric
compounds is that they can rotate the polarization plane of the plane-polarized light. As
shown in Figure 21.11, the electromagnetic radiation of the plane-polarized light shows
oscillations in only two directions in one plane.
Introduction to Pharmaceutical Chemical Analysis
Figure 21.10 Visualization of the chirality concept
354
Chemical Analysis of Drug Substances
355
Figure 21.11 Instrument for measurement of optical rotation
Some enantiomers rotate the plane-polarized light to the right and this is highlighted by
placing (þ) before the name of the compound (or d for dextrorotatory). The other mirror
image rotates the plane-polarized light as much to the left and this is marked with () or l
in front of the name (l for levorotatory). If a compound contains an equal amount of (þ) and
() enantiomers it is called a racemic mixture or a racemate, and as the two enantiomers
rotate the plane-polarized light equally to the right and left, respectively, the result is no
rotation of the plane-polarized light. Since optical rotation is the only physical property that
is different for enantiomers it is often used for characterization and identification of chiral
pharmaceutical raw materials. Optical rotation is measured with a polarimeter, and the
principle of this instrument is shown in Figure 21.11.
The light source in the instrument is a sodium lamp which emits light at 589.3 nm. Light
passes through a fixed Nicol prism (polarization lense) that plane-polarizes the light. The
plane-polarized light passes through a tube containing the sample solution. If the
compound is a liquid it can be filled directly into the tube, and if the compound is a
solid it has to be dissolved before placing it in the tube. At the other end of the tube a
second Nicol prism that can be rotated is placed, and the operator observes the light
intensity through this prism. If there is no solution in the polarimeter tube (or if there is no
optically active solution), a maximum of light will pass through when the polarization
planes of the two prisms are aligned. When the Nicol prism at the observer is rotated, the
observed light intensity is reduced because the polarization planes are out of alingment. At
90 rotation the operator will not observe any light transmission. When an optically active
sample is introduced into the polarimeter tube the plane of the polarized light is rotated,
and the operator observes a sharp reduction in light intensity in relation to the starting
point where the polarization planes of the two prisms are aligned. The operator must rotate
the prism either right or left to observe the maximum light transmission. The angle
expressed in degrees ( ) that the prism must be rotated to again observe the maximum light
transmission can be read accurately on a scale divided in degrees and the angle is called
the observed rotation.
The optical rotation observed is a result of the compound that are in the polarimeter tube.
Therefore, the observed rotation depends on: (i) the compound measured, (ii) the length of
the polarimeter tube, (iii) the concentration of the compound, (iv) the solvent used for the
356
Introduction to Pharmaceutical Chemical Analysis
sample preparation and (v) the temperature of the sample. For identification purposes it is
convenient to eliminate the contributions (ii), (iii), (iv) and (v) and only operate with a
parameter that depends on the structure of the compound, and therefore the optical rotation
should always be calculated as the specific optical rotation ½a20
D defined as follows in the
pharmacopoeias for solutions of solids and undiluted liquids, respectively:
½a20
D ¼
1000 a
lc
½a20
D ¼
a
l r20
For a solution of solids
For neat liquids
ð21:1Þ
ð21:2Þ
where a is the observed rotation (degrees), l is the length of the polarimeter tube (dm), c is
the concentration of the compound (g/l) and r20 is the density of the liquid at 20 C (g/cm3).
D indicates that the measurement is done using a sodium lamp at 589.3 nm, as this is
traditionally termed the D line.
Raw materials that are liquids at room temperature is usually measured directly without
dilution, while solids must be dissolved before they can be examined by polarimetry. Raw
materials are typically dissolved either in water, dilute HCl, dilute NaOH, methanol,
ethanol, acetonitrile or methylene chloride, and the concentration of the compound in the
sample solution for polarimetry is typically in the range 10–100 mg/ml depending on the
specific rotation. Concentrated test solutions are an advantage if the specific optical rotation
of the material is small in order to ensure a sufficiently reliable figure of the measured value
of the observed rotation. The specific optical roration can be measured at other wavelengths
(l) and other temperatures (t) but this should then be stated as ½atl .
In the pharmacopoeias the optical rotation is given without any units, but in fact the unit is
degree milliliters per decimeter gram, which is understood.
Using a polarimeter the observed rotation of a solution of the compound is measured if it
is a solid or directly on the compound if it is a neat liquid. From the observed rotation the
specific optical rotation is calculated, and this value is compared with the requirement in Ph.
Eur. When the limit for specific optical rotation is based on dried substance/anhydrous
substance the calculated value has to be corrected for the value obtained for loss on drying/
water. The requirement for specific optical rotation is stated as a range. If the value found
falls within the claimed range the identity test is positive. An example of identification of
simvastatin using polarimetry is shown in Box 21.5. For this drug substance polarimetry is
used as an identification procedure for the “First identification” in the Ph.Eur.
Polarimetry is the simplest technique available to differentiate between the different
enantiomeric forms of the same compound. It is normally used to characterize drug
substances having optical rotation or to verify if the substance is a racemic mixture.
The technique is relatively unspecific as a number of different compounds may have about
the same value for specific optical rotation. Therefore, the method cannot stand alone as an
identification test, but must be combined with other methods.
21.4.2.6 High Performance Liquid Chromatography
Earlier in this chapter, the use of TLC for identification of pharmaceutical raw materials was
discussed. TLC is the simplest chromatographic technique used for control of raw materials.
With the intensified use of HPLC for test for related substances this technique has also been
Chemical Analysis of Drug Substances
357
Box 21.5 Identification of simvastatin by polarimetry according to Ph.
Eur. 7th edn (2011)
Specific optical rotation (2.2.7): þ 285 to þ 300 (dried substance).
Dissolve 0.125 g in acetonitrile R and dilute to 25.0 ml with the same solvent.
Discussion of the procedure:
0.125 g of the compound is dissolved in acetonitrile and diluted to 25.0 ml with
acetonitrile. The observed rotation is measured and the specific optical rotation is
calculated.
The specific optical rotation should be within the range þ 285 to þ 300 (dried
substance).
Calculation Example:
An observed rotation of þ2.9 is measured using a polarimeter tube of 20 cm. The
concentration of the compound in the sample solution is:
c ¼ 0:125 g=0:025 l ¼ 5:0 g=l
The specific optical rotation is calculated to be:
½a20
D ¼ 1000 a=ðl cÞ ¼ 1000 2:9=ð2 5Þ ¼ þ290
The maximum value for loss on drying is 0.5% and only 0.3% is found in this test. Thus
the compound contains 99.7%.
þ290 100/99.7 ¼ þ290.9 and this is within the requirement prescribed in the
monograph in Ph.Eur.
The reference (2.2.7) means that the general monograph on optical rotation in Ph.Eur.
where general informations on optical rotation can be found has to be consulted.
introduced as an identification parameter. Liquid chromatography is described in Chapter
13. When using HPLC for identification it is common to refer to an HPLC method described
under “Tests” (i.e., related substances) or under “Assay” in the monograph where first a
solution of the reference substance and then a solution of the test compound are injected into
the HPLC. The requirement for a positive identification is that the principal peak in the
chromatogram obtained with the test solution is similar in retention time and size to the
principal peak in the chromatogram obtained with the reference solution. This formulation
of the requirement is to ensure that only the main peaks of the two solutions are considered.
An example of the use of HPLC for identification of a drug substance is shown in Box 21.6
for calcitriol, which is a vitamin D analog.
HPLC provides higher separation efficiency than TLC. In principle HPLC could replace
TLC in the identification, but HPLC requires a far more expensive equipment which is not
common to have in pharmacies and similar laboratories. Therefore TLC is kept for
identification in many monographs where HPLC has been introduces for test for related
358
Introduction to Pharmaceutical Chemical Analysis
Box 21.6 Identification of calcitriol according to Ph.Eur. 7th edn (2011)
by HPLC
Examine the chromatograms obtained in the assay.
Discussion of the procedure:
A test solution prepared by dissolving 1.0 mg of raw material in 10.0 ml of the mobile
phase. A reference solution prepared similarly by dissolving 1.000 mg of calcitriol CRS
in 10.0 ml of mobile phase. 50 ml of each of the solutions injected separately in a HPLC
apparatus consisting of the following main components:
Mobile phase: 1.0 ml/min consisting of 45 vol of a 1.0 g/l Tris (hydroxy-methyl)
aminomethane (pH 7.0–7.5) with phosphoric acid and 55 vol% of acetonitrile.
Column: 25 cm long, 4.6 mm inner diameter, packed with 5 mm silica particles modified
with C8H17 groups.
Retention time
min.
Chromatogram of the chemical reference substance
Detector signal
Detector signal
Detection: UV spectrophotometry at 230 nm.
Retention time
min.
Chromatogram of the sustance to be tested
Requirements for positive identification:
Retention time and peak area for the main peak in the chromatogram of the raw material
(test solution) should be similar to the corresponding data in the chromatogram for the
reference standard (reference solution).
Comments on the procedure:
Calcitriol is a relatively hydrophobic compound, and therefore a C8 column is chosen in
order not to obtain too strong retention. Isocratic elution is used since the assay where
only the main peaks are considered do not require sophisticated separation. Calcitriol is a
very expensive compound used in very low doses. This is the reason for only using 1.0 mg
for the solutions. The substances is dissolved in the mobile phase as this is a suitable
solvent for calcitriol, and as this is standard procedure in HPLC to avoid a sample
solution with a higher elution strength than the mobile phase.
Chemical Analysis of Drug Substances
359
substances and also for identification or assay. There is however reasons to believe that
HPLC will be used more than TLC in the future.
21.4.2.7 Chloride and Sulfate Identification
As previously mentioned, IR spectrophotometry is the most important techniques for the
identity testing of raw materials. A weakness of the IR spectrophotometry is that it is less
suitable for identification of inorganic ions. This is very relevant for a variety of APIs because
they are typically made available as hydrochlorides or sulfates. However, even though the
difference in chemical composition is minor between an organic base and its hydrochloride
and sulfate salts major differences in the IR spectra of such compounds are normally very
distinct. An example of this is shown in Figure 21.12 for morphine and its hydrochloride and
sulfate salts.
This means that although most of the fine structure in the IR spectrum is coming from the
“organic part” of the salt the different salt can be distinguished by IR spectrophotometry. In
order to avoid mistakes, however, the anions are also identified. The identity reactions for
the inorganic counter ions are described in Ph.Eur. (Chapter 1). The chloride ion in
hydrochlorides of active substances is normally detected in a solution of the API by adding
a solution of silver nitrate. This is illustrated in Box 21.7. If the API is a hydrochloride a
white precipitate of silver chloride is observed.
Sulfate salts of drugs can be detected by addition of barium chloride as shown in Box 21.8.
21.5 How to Test for Impurities in Pharmaceutical Raw Materials
As mentioned in Section 21.2 almost all monographs contain a section called “Tests” where
methods for purity testing are given.
The purpose of these methods is to ensure that the raw materials do not contain
undesirable contaminants of various organic or inorganic compounds. The purity tests are
often qualitative methods or limit tests, which means that they are intended to control that
the content of possible contaminants does not exceed a certain limit. In addition, some
tests are quantitative and are used to quantify the content of a given impurity. Each
monograph contains typically 3–12 different tests. These are selected to identify and limit
the main impurities that are known for the raw material. It is primarily a question of
qualified organic impurities, volatile contaminants (solvents and water) and inorganic
impurities. Such a limited range of purity tests will obviously not be able to detect all
possible impurities. Too extensive testing is not desirable, as this could initiate further
purification of pharmaceutical raw materials beyond what is medically relevant. The main
procedures for purity testing is reviewed below. In addition to these, there are a large
number of other tests that are more or less drug-dependent. This latter group is not
discussed here.
21.5.1
Main Purity Tests for Well Defined Chemical Compounds
The main methods for purity testing of the raw materials of the type well defined chemical
compounds are summarized in Table 21.10.
360
Introduction to Pharmaceutical Chemical Analysis
Figure 21.12 IR spectra of morphine base (top), morphine hydrochloride (middle) and
morphine sulfate (bottom)
Chemical Analysis of Drug Substances
361
Box 21.7 Principle of the identity reaction for chlorides
To ensure that the API is a hydrochloride an aqueous solution of the raw material is
prepared. To this solution an aqueous solution of silver nitrate is added. If the compound is
a hydrochloride, chloride ions will be present in the solution and white silver chloride will
precipitate because the low solubility product (1.8 1010) of the silver chloride salt.
þ
ClðaqÞ
þ AgðaqÞ
! AgClðsÞ # ðprecipitateÞ
To further ensure that it is silver chloride that has been precipitated, the precipitate is
isolated and washed with water. Then the precipitate is suspended in water and concentrated ammonia is added. The precipitate of silver chloride will then be redissolved, which
is necessary for the identity test to be positive. The precipitate of silver chloride is
redissolved as the silver ions form a very stable, soluble complex with ammonia:
þ
AgðaqÞ
þ 2NH3ðaqÞ ! Ag ðNH3 Þ2þ
Box 21.8 Principle of the identity reaction for the detection of sulfates
To ensure that the API is present as a sulfate an aqueous solution of the raw material is
prepared and an aqueous solution of barium chloride is added. If the raw material is a
sulfate, sulfate ions are present in the solution and will form a white precipitate
consisting of the barium sulfate salt having a very low solubility (1.1 1010):
þ
SO24 þ Ba2ðaqÞ
! BaSO4ðsÞ # ðprecipitateÞ
An example of how IR spectrophotometry combined with a simple chemical reaction for
identification are shown in Box 21.9 for the antihistamine drug chlorcyclizine hydrochloride. The chlorcyclizine ion of the raw material is detected by IR spectrophotometry,
while the chloride ion is detected by precipitation with silver nitrate.
21.5.1.1 Appearance of Solution
In many monographs a test for appearance of solution is included as the first test. The raw
material is dissolved in a specified solvent, which typically can be water, dilute HCl, dilute
NaOH, methanol, ethanol, dichloromethane or dimethylformamide. The concentration of
the raw material in the sample solution is high and is typically in the range 10–200 mg/ml. If
the solution is prescribed to be clear it must have a clarity similar to water. If some
opalescence is allowed, the test solution is compared to a well defined suspension made from
a mixture of hydrazine sulfate and hexamethylenetetramine. In this way traces of impurities
having a low solubility in the selected solvent are detected. Furthermore, the color of
solution is evaluated. Most raw materials of the type well defined chemical compounds are
colorless, while a number of degradation products can be colored. This means that the test
can be quite effective in detecting certain types of quality problems despite the fact that it is a
very simple test. To ensure an objective assessment of the clarity or color the sample solution
362
Introduction to Pharmaceutical Chemical Analysis
Box 21.9 Identification of chlorcyclizine hydrochloride according to Ph.
Eur. 7th edn (2011)
The monograph for chlorcyclizine hydrochloride in Ph.Eur. prescribe:
First identification: B, D.
B: Examine by infrared absorption spectrometry (2.2.24), comparing with the spectrum
obtained with chlorcyclizine hydrochloride CRS. Examine the substances prepared
as discs.
D: It gives reaction (a) of chlorides (2.3.1).
Main features of the first ID procedures:
IR spectroscopy of the raw material is performed in the same manner as discussed in
Box 21.1. The test for chloride is described in Ph.Eur. paragraph (2.3.1) and this includes
the use of raw material equivalent to 2 mg Cl (19 mg of the substance having
a molecular weight of 337.3 – please check this yourself!) dissolved in 2 ml of water.
This solution is acidified with dilute nitric acid and 0.4 ml of silver nitrate solution
R1 (42.5 g/l) is added. Shake and allow to stand. A curdled, white precipitate is formed.
Centrifuge and wash the precipitate three times, each time with 1 ml of water. This has to
be performed rapidly in subdued light. The precipitate is suspended in 2 ml of water and
1.5 ml of ammonia R (17.5 g/l). The precipitate will then redissolve and the test is
considered to be positive.
is compared to a comparison sample. If the sample solution is more turbid than the
comparison sample or if the sample solution is more colored than the comparison sample,
the compound fails the test and therefore it does not meet the quality requirements given in the
monograph. Box 21.10 presents an example of a test for appearance of solution of ibuprofen.
The preparation of the sample solution is often prescribed separately above the test for
appearance of solution as this sample solution often is used in some of the other purity tests.
Table 21.10 Overview of the most important tests for purity testing of well
defined chemical compounds
Method for purity testing
Impurities to be controlled
Appearance of solution
Absorbance
Acidity/alkalinity
Specific optical rotation
Related substances
Specific impurities
Residual solvents
Inorganic anions
Heavy metals
Sulfated ash
Loss on drying
Water
Particles and colored impurities
UV-absorbing impurities
Acid and alkaline impurities
Optically active impurities
Related organic impurities
Each specific impurity
Organic solvents
Unfamiliar anions
Heavy metals
Inorganic material
Volatile substances
Water
Chemical Analysis of Drug Substances
363
Box 21.10 Test for appearance of solution of ibuprofen according to Ph.
Eur. 7th edn (2011)
Solution S: Dissolve 2.0 g in methanol R and dilute to 20 ml with the same solvent.
Appearance of solution: Solution S is clear (2.2.1) and colorless (2.2.2, Method II).
Discussion of the procedure:
A sample solution of ibuprofen is prepared by dissolving 2.0 g of raw material in
methanol and diluting this to 20 ml. The clarity and color of this solution is evaluated
against a comparison sample of water. The sample solution and the comparison sample
solution are placed into two test tubes, which shall have a diameter of between 15 and
25 mm and which shall be made of clear, colorless glass. The solutions are filled to a
height of 40 mm, and the comparison has to be made in daylight vertically through the
sample tubes. In assessing the clarity the observation has to be made against a black
background, while assessment of the color should be done against a white background.
Viewing
Test
solution
Viewing
Reference
solution
Clarity and degree of
opalescence of liquids
Test
solution
Reference
solution
Degree of coloration
of liquids
Requirements for a positive test:
Sample solution must not be more turbid or contain particles or be more colored than
water.
Detailed information on how to perform the test for clarity and degree of opalescence of
liquids (Chapter 2.2.1) and degree of coloration of liquids (Chapter 2.2.2) are found in the
chapters given in parentheses.
Some compounds when dissolved give a somewhat unclear or colored solution despite the
fact that the level of impurities is still very low. In such cases water is not used as the
comparison sample. In Chapters 2.2.1 and 2.2.2 details of the preparation of comparison
sample solutions used for compounds that give some opalescence and/or color of solution
when dissolved are given.
In monographs on newer organic compounds not used for parenteral applications the test
for appearance of solution is often deleted. This is primarily due to the fact that the HPLC
test for related substances is supposed to be able to detect any significant colored impurity
and normally organic chemicals do not contain impurities that do not dissolve.
364
Introduction to Pharmaceutical Chemical Analysis
21.5.1.2 Absorbance
The test for absorbance is identical to the test used for identification called “Ultraviolet and
visible absorption spectrophotometry” as described in Section 21.4.2.2. Measurement of
absorbance in the UV-Vis range by using a spectrophotometer (Chapter 7) is prescribed as a
purity test in a number of monographs. A relatively concentrated solution of the raw material
(10–100 mg/ml) is typically used prepared in dilute HCl, dilute NaOH, methanol, ethanol,
dichloromethane or dimethylformamide. Normally the absorbance is measured at a
wavelength where one or more potential contaminants have strong UV absorbance and
where the main substance has no or little UV absorbance. The reported UV absorbance is in
the monograph typically limited to a given absorbance value. The test is valuable for
detection of impurities that do not have their UVabsorbance in the same wavelength range as
the main compound. An example of purity testing uisng absorbance measurement is shown
in Box 21.11.
Box 21.11 Test for absorbance prescribed in the monograph for
esomeprazole magnesium trihydrate according to Ph.Eur. 7th edn (2011)
Absorbance (2.2.25): maximum 0.20 at 440 nm.
Dissolve 0.500 g in methanol R and dilute to 25.0 ml with the same solvent. Filter the
solution through a membrane filter (nominal pore size 0.45 mm).
Discussion of the procedure:
A solution of the compound to be tested is prepared by dissolving 0.500 g of the
compound to 25.0 ml of methanol. The absorbance of this solution is measured at
440 nm, and the value should not exceed 0.20. This corresponds to a maximum level of
impurity F or G (see figure below) of about 0.04% (w/w).
H3CO
N
O
N
CH3
Omeprazole
N
H3CO
N
Impurity F
O
H3CO
N
CH3
S
O
OCH 3
S
CH3
S
CH3
NH
N
CH3
N
N
CH3
Impurity G
Comments on the procedure:
The raw material is dissolved in methanol because this is a good solvent for esomeprazole, and because methanol has no absorbance itself at 440 nm. The number of digits
indicates the precision required in this method. The solution being tested is highly
concentrated, and therefore will be able to detect even very small impurities (0.035% w/w)
at 440 nm where esomeprazole does not absorb, but where impurity F and G has
absorbance. To avoid false high absorbance due to particles the sample solution is filtered
through a 0.45 mm filter.
Chemical Analysis of Drug Substances
365
21.5.1.3 Acidity/Alkalinity
Measurement of acidity/alkalinity is a purity test which also find relatively large
application in the Ph.Eur. In this test the content of protolytic impurities, that is, impurities
having acid or base properties that are different from those of the main compound are
evaluated. The test is normally carried out in one of two ways. Either pH is measured
directly in a solution of the compound using a pH meter, or alternatively a semiquantitative titration is carried out. In the latter procedure an aqueous solution of the
compound is made and is added an acid–base color indicator. Then a certain amount of
acid or base is added to make the pH lower or higher, and the color change in the solution
due to the pH change should be observed. After this, either a base or an acid is added to
adjust pH back into the pH area from the start of the procedure, and a new color change
should be observed. The acid and base are added in small quantities just sufficient to
change the pH and cause color shifts. If the raw material contains significant amount of
protolytic impurities they can affect the pH of the sample solution, and the characteristic
color shifts may not occur or may be changed. In cases where the solution of the compound
has a certain buffer capacity pH measurements are preferred. For ingredients that do not
have the buffer capacity a semi-quantitative titration is used. Boxes 21.12 and 21.13 show
examples of testing dopamine and fenoterol for acidity/alkalinity.
21.5.1.4 Optical Rotation
Under the section on identification (Section 21.4.2), it was discussed that the measurement
of optical rotation in some cases is used for identification purposes. The use of this technique
is even more prevalent during purity testing, and it is used to determine whether (i) an
optically active raw material has sufficiently high purity or (ii) an optically inactive
raw material contains traces of optically active impurities. The name of the test is
“Specific optical rotation”, and the value is calculated from an observed rotation (see
Section 21.4.2.5). In some cases the observed rotation is used directly, and the test is then
called “Angle of optical rotation,” or “Optical rotation”. The principle is to prepare a
Box 21.12 Test for acidity or alkalinity in dopamine hydrochloride
according to Ph.Eur. 7th edn (2011)
Acidity or Alkalinity. Dissolve 0.5 g in carbon dioxide-free water R and dilute 10 ml with
the same solvent. Add 0.1 ml of methyl red solution R and 0.75 ml of 0.01 M sodium
hydroxide. The solution is yellow. Add 1.5 ml of 0.01 M hydrochloric acid. The solution
is red.
Main features of the procedure:
0.5 g raw material is dissolved in carbon dioxide-free water and diluted to 10 ml with the
same solvent. 0.1 ml of a solution of the pH indicator methyl red and 0.75 ml of 0.01 M
NaOH are added. The solution should be yellow, which occurs for methyl red at
pH > 6.2. When 1.5 ml of 0.01 M HCl are added the solution should become red, which
happens to methyl red at pH < 4.4. The solvent has to be carbon dioxide-free to eliminate
the buffering capacity of dissolved carbon dioxide.
366
Introduction to Pharmaceutical Chemical Analysis
Box 21.13 Test for pH in fenoterol hydrobromide according to Ph.Eur.
7th edn (2011)
Solution S: Dissolve 2.00 g in carbon dioxide-free water R and dilute two 50.0 ml with
the same solvent.
pH (2.2.3): 4.2–5.2 for solution S.
Main features of the procedure:
2.00 g of raw material is dissolved in carbon dioxide-free water and diluted to 50.0 ml with
the same solvent. pH is measured using a pH meter and shall bewithin the range pH 4.2–5.2.
Section (2.2.3) in Ph.Eur. defines how to measure the pH, the requirements for the pH
meter and how it shall be calibrated to provide accurate pH values. Carbon dioxide-free
water is freshly boiled water R cooled to room temperature.
solution of the raw material on which the optical rotation is measured using a polarimeter
(see Section 21.4.2.5). From the optical rotation the specific optical rotation is calculated,
and then assessed whether it falls within the specified range listed in the Ph.Eur.
monograph. The raw materials are typically dissolved either in water, weak HCl,
weak NaOH, methanol, ethanol, acetonitrile or methylene chloride, and the concentration
of raw material in the sample solution for polarimetry is typically in the range 10–100 mg/
ml. An example of the measurement of optical rotation of an optically active raw material
is shown in Box 21.5. This procedure, which was discussed during the identification of
simvastatin, is also used for purity testing and is intended to determine whether
simvastatin contains optically active or inactive impurities. It is common that tests for
purity are also used for identification when they serves both purposes. Similarly, an
example of ibuprofen is shown in Box 21.14, where an optical inactive raw material (a
racemate) is tested for optically active impurities.
21.5.1.5 Related Substances
Avery important purity test that is included in almost all monographs for organic substances is
a test for the content of related substances, that is, organic impurities. It is a test for impurities
Box 21.14 Test for optical rotation of ibuprofen according to Ph.Eur. 7th
edn (2011)
Optical rotation (2.2.7): 0.05 to þ 0.05 . Dissolve 0.50 g in methanol R and dilute to
20.0 ml with the same solvent.
Discussion of the procedure:
0.50 g of raw material is dissolved in methanol and diluted to 20.0 ml with the same
solvent. The optical rotation should be within 0.05 to þ0.05 . This test also
distinguishes the racemate from the optically active S-enantiomer which is also
commercially available.
Chemical Analysis of Drug Substances
367
Box 21.15 Possible impurities of type related substances in paracetamol
Cl
HO
O
O
N
H
OH
O
N
H
H 3C
H 3C
HO
O2N
NOH
O
OH
HO
OH
OH
Cl
O
N
H
N
H
O
N
H
N
H
OCOCH3
O
OH
OH
H2N
O
similar in structure to the main compound, but also impurities with other structures
originating from starting materials in the synthesis of the compound can be included.
Examples of contaminants on the type of related compounds are shown in Box 21.15 for
paracetamol. The test is called “Related substances.” Typically, the analytical chemical
technique used is either HPLC (see Chapter 13) or TLC (see Chapter 12), but in some cases it
may also be appropriate to use GC (see Chapter 14). GC is used if the relevant impurities are of
relativelyhigh volatility, in all other cases either HPLC or TLC are used. In anycase, a solution
of the raw material is needed with a relatively high concentration to be able to detect very low
amount of the impurities (corresponding to less than 0.05% of the main compound). All three
analytical chemical techniques mentioned are separation methods, and this is needed to be
able to separate the related impurities from the main substance in the raw material.
Box 21.16 shows how HPLC is used in the test for related substances in omeprazole.
Box 21.16 Test for related substances in the omeprazole according to Ph.
Eur. 7th edn (2011)
Related Substances: Liquid chromatography (2.2.29). Prepare the solutions immediately before use.
Test solution. Dissolve 3 mg of the substance to be examined in the mobile phase and
dilute to 25.0 ml with the mobile phase.
. Reference solution (a). Dissolve 1 mg of omeprazole CRS and 1 mg of omeprazole
impurity D CRS in the mobile phase and dilute to 10.0 ml with the mobile phase.
. Reference solution (b). Dilute 1.0 ml of the test solution to 100.0 ml with the mobile
phase. Dilute 1.0 ml of this solution to 10.0 ml with the mobile phase.
. Reference solution (c). Dissolve 3 mg omeprazole for peak identification CRS (containing impurity E) in the mobile phase and dilute to 20.0 ml with the mobile phase.
368
Introduction to Pharmaceutical Chemical Analysis
Column:
. Size: l ¼ 0.125 m, Ø ¼ 4.6 mm;
. Stationary phase: octylsilyl silica gel for chromatography R (5 mm).
Mobile phase: mix 27 vol of acetonitrile R and 73 vol of a 1.4 g/l solution of disodium
hydrogen phosphate R previously adjusted to pH 7.6 with phosphoric acid R.
Flow rate: 1 ml/min.
Detection: spectrophotometer at 280 nm.
Injection: 40 ml.
Run time: 5 the retention time of omeprazole.
Omeprazol
Unknown
impurity
Unknown
impurity
Impurity D
Impurity E
Test solution
t2
Rs = (t2 - t1)/0.5 * (tw1 + tw2)
t1
tw1
Reference solution (a)
tw2
Reference solution (b)
Reference solution (c)
H3C
CH3
NH
OCH3
N
S
O
O
N
CH3
Omeprazole impurity D
Identification of impurities: use the chromatogram obtained with reference solution (a)
to identify the peak due to impurity D; use the chromatogram supplied with omeprazole
for peak identification CRS and the chromatogram obtained with reference solution (c)
to identify the peak due to impurity E.
Chemical Analysis of Drug Substances
369
Relative retention: with reference to omeprazole (retention time ¼ about 9 min): impurity E ¼ about 0.6; impurity D ¼ about 0.8.
System suitability: reference solution (a).
Resolution: minimum 3.0 between the peak due to impurity D and omeprazole; if
necessary, adjust the pH of the aqueous part of the mobile phase or its proportion of
acetonitrile; an increase in pH improves the resolution.
Limits:
. Impurities D, E: for each impurity, maximum 0.1%.
. Unspecified impurities: for each impurity, maximum 0.10%.
. Total: maximum 0.5%.
. Disregard limit: half the area of the principal peak in the chromatogram obtained
with reference solution (c), 0.05%.
Comments on the procedure:
The raw material and reference solutions are dissolved in mobile phase as this is a
suitable solvent for the compounds, and because the mobile phase is a very suitable
medium for injection into the HPLC. The column is a C8 column, which means that
the separation principle is reversed phase chromatography. The C8 stationary phase
together with a mobile phase containing 27% of acetonitrile provides a suitable
retention of the analytes. The test and reference solutions are made up in the mobile
phase and as omeprazole has a limited stability the solutions have to be prepared
immediately before use. The pH of the mobile phase is adjusted to a pH where
omeprazole has a fair solubility and a fair stability. Omeprazole is a weak acid (pKa
approx. 7.9) and for all ionizable substances it is important to have pH control in the
mobile phase to ensure stable retention times. Normally substances are separated at a
pH significantly different from the pKa value. In this example, this is not the case
since a phosphate buffer at pH 7.6 is selected. This is because the substances are
separated better at pH 7.6 and because omeprazole is unstable in acidic solution. A
system suitability test has to be performed before analysis of the sample can take
place. If the system suitability requirement of a minimum resolution of 3.0 between
impurity D and omeprazole cannot be met, advice is given on how to obtain this by
modification of the mobile phase. A decrease in the acetonitrile concentration will
increase the retention time but it is not known if this will provide the sufficient
separation. An increase in pH is reported to improve the resolution and this is
probably the way to go if optimization is needed. A standard flow rate of 1 ml/min is
used and a run time of 5 the retention of omeprazole is needed to elute all the
possible impurities. A run time is always prescribed when an isocratic HPLC system
is used in order to inform the user about the time needed to elute all possible
impurities from the column. The injection volume (40 ml) is a little above the usual
volume but is still acceptable with a column having an internal diameter of 4.6 mm.
Omeprazole and its impurities are detected at 280 nm, where all relevant compounds
have relatively high UV absorption.
370
Introduction to Pharmaceutical Chemical Analysis
Acceptance criteria:
No peaks in the chromatogram of sample solution (see below), apart from the peaks
corresponding to omepazole and impurity D and E, will be larger in area than the top of
omeprazole in the highly diluted reference solution (b). This means that peaks with areas
larger than the peak corresponding to a 0.10% dilution of omeprazole are not accepted.
Assuming that the UVabsorbance of omeprazole and the impurities are similar it means
that no single impurity can be accepted in an amount larger than 0.10% (w/w). Impurity
D and E are limited to 0.1% which in practice means that they can be present in amounts
up to 0.144% due to downwards rounding of this figure. In total impurity D, E and other
impurities present above the disregard limit 0.05% should be maximum 0.5%.
In addition, the resolution, RS, (see also Chapter 10) between omeprazole and impurity
D in the reference solution (a) should be at least 3. This is to ensure that the HPLC system
has sufficient separation capability to separate the different substances in a satisfactory
manner. Reference solution (c) is used for the identification of impurity E.
If the separation selectivity in the HPLC system is inadequate some impurities
may co-elute resulting in an impurity peak that is too large to comply with the
limit. Alternatively, a serious impurity can co-elute with the peak for omeprazole
and will therefore not be detected. The raw material can thus erroneously be released
for production.
Although HPLC is replacing TLC in test for related substances TLC may still be used to
test for certain impurities which are difficult to analyze or detect by GC or HPLC.
Box 21.17 shows how TLC is used for a test for a related substance in amiodarone
hydrochloride.
Box 21.17 Test for impurity H in related substances in amiodarone
hydrochloride according to Ph.Eur. 7th edn (2011)
Impurity H: Thin-layer chromatography (2.2.27). Prepare the solution immediately
before use and keep protected from bright light.
Test solution: Dissolve 0.500 g of the substance to be examined in methylene chloride R
and dilute to 5.0 ml with the same solvent.
Reference solution (a): Dissolve 10.0 mg of (2-chloroethyl)diethylamine hydrochloride
R (impurity H) in methylene chloride R and dilute to 50.0 ml with the same solvent.
Dilute 2.0 ml of this solution to 20.0 ml with methylene chloride R.
Reference solution (b): Mix 2.0 ml of the test solution and 2.0 ml of reference
solution (a).
Plate: TLC silica gel F254 plate R.
Mobile phase: anhydrous formic acid R, methanol R, methylene chloride R (5 : 10 : 85,
v/v/v).
Chemical Analysis of Drug Substances
371
Application: 50 ml of the test solution and reference solution (a); 100 ml of reference
solution (b).
Development: two-thirds of the plate.
Drying: in a current of cold air.
Detection: spray with potassium iodobismuthare solution R1 and then with dilute
hydrogen peroxide solution R; examine immediately in daylight.
System suitability: reference solution (b) – the spot due to impurity H is clearly visible.
Solvent front
O
H3 C
CH3
I
O
N
CH3
O
1. Test solution
I
2. Reference solution (a)
Amiodarone
3. Reference solution (b)
CH3
N
•
•
•
CH3
Cl
Impurity H = 2-chlorotriethy lamine
1
2
3
Limit for impurity H: any spot with the same Rf as the spot due to impurity H in the
chromatogram obtained with the test solution, is not more intense than the spot in the
chromatogram obtained with reference solution (a) (0.02%).
The test for related substances generally does not allow unknown related substances to be
present in the raw material in quantities exceeding 0.10% (w/w). Previously this limit was
0.1% (w/w) which using rounding in practice could correspond to a limit upto 0.144%.
Impurities present in concentrations above 0.10% are to be identified, and if the concentration exceeds 0.15% the impurity also has to be qualified which means its toxicity have to be
known. In most cases these limits represent sufficient demands seen from a medical
perspective. Impurities below 0.10% (w/w) in the raw material will result in a very low
daily intake of impurities when using the finished product. For drug substances given in
higher doses over a longer period of time lower limits for the impurities are valid. In some
cases, however, it is known that some impurity can have a significant toxic effect, and for
such contaminants much stricter limits will apply. In pethidine hydrochloride used for
parenteral administration a limit for an impurity called methyl-phenyl-tetrahydropyridine
(MPTP) is as low as 0.1 ppm (one part in 10 million).
If the test for related substances is not adequate to detect such low concentrations of toxic
impurities, separate test for those are given in the monographs similar to what was shown for
impurity H in amiodarone hydrochloride.
372
Introduction to Pharmaceutical Chemical Analysis
21.5.1.6 Solvent Residues
Pharmaceutical raw materials can contain residual solvents used in the synthesis or in the
purification process. The content of the solvents must be low to prevent substantial intake in
patients using the final pharmaceutical preparations. As discussed previously, the solvents in
Class 1 (see Table 21.7) should not be present in APIs. The content of Class 2 solvents
permitted in APIs is calculated on basis of a so called “permitted daily exposure” (PDE)
value for the individual solvent:
Concentration ðppmÞ ¼
1000 PDE
dose
ð21:3Þ
where PDE is in mg/day and dose is given in g/day.
The PDE as well as the concentration limit for each Class 2 solvent is given in the
pharmacopoeia.
Class 3 solvents are typically limited to a maximum of 0.5% (w/w) determined by loss on
drying. Limits for the individual solvents are given in the pharmacopoeia.
The requirement to the presence of residual solvents in pharmaceutical raw materials is a
general demand put forward in the general monograph on “substances for pharmaceutical
use”. All pharmaceutical raw materials shall fulfil these requirements even if it is not stated
in the respective monographs.
The control of residual solvents is performed using static headspace gas chromatography
as described in the pharmacopoeia. The principle is that a small amount of raw material is
dissolved in a suitable solvent (e.g., water or N,N-dimethylformamide) and placed in an
airtight glass vial. The vial is equilibrated at a certain temperature for a given time to obtain
an equilibrium of the residual solvent in the headspace in the vial (the air above the liquid
phase) and the amount of residual solvent in the air is proportional to the amount of residual
solvent in the substance. A small sample of air in the vial is sampled with a syringe and is
injected into a GC system.
The various components are separated from each other in the GC column, and are
detected using an FID detector. The amount of solvent is determined quantitatively using
calibration standards of solvents prepared similarly to the sample or using standard addition.
When a higher limit for a given solvent (higher than the one given in the general
monograph on residual solvents) is allowed, a separate test appears in the monograph for the
raw material. The test has the name of the solvent to be tested.
An example of a test for residual solvent by GC is shown in Box 21.18 for omeprazole.
21.5.1.7 Foreign Anions
A series of monographs gives tests for the presence of specific foreign anions. This is
because acids are often used during the synthesis of the raw materials, and traces of these can
remain in the raw material. For inorganic compounds produced from inorganic materials
tests for a number of different foreign anions are often prescribed. In monographs for
organic compounds it is limited to one or two tests, typical test for trace amounts of chloride
(the test is called “Chlorides”) and trace amounts of sulfate (the test is called “Sulfates”). For
organic compounds it is thus not normal to test for many different foreign anions even
though they theoretically could be present. The test for chloride involve the raw material to
be dissolved in a suitable aqueous solvent (often containing dilute nitric acid), and then
Chemical Analysis of Drug Substances
373
Box 21.18 Test for chloroform and methylene chloride in omeprazole
according to Ph.Eur. 7th edn (2011)
Main features of the procedure:
Headspace gas chromatography (2.2.28): use the standard additions method.
Test solution: Place 0.50 g of the substance to be examined in a 10-ml vial. Add 4.0 ml
dimetylacetamide R and stopper the vial.
Column:
. Material: fused silica;
. Size: l ¼ 30 m, Ø ¼ 0.32 mm;
. Stationary phase: cross-linked poly[(cyanopropyl)-(phenyl)][dimethyl]siloxane R
(film thickness 1.8 mm).
Carrier gas: nitrogen for chromatography R.
Static headspace conditions that may be used:
. Equilibration temperature: 80 C;
. Equilibration time: 1 h.
Detection: flame ionization.
Limits:
. Methylene chloride: maximum 100 ppm;
. Chloroform: maximum 50 ppm.
Calculation example:
In the monograph for gas chromatography (2.2.28) in the section for static headspace GC it
is stated that at set of containers should be used. All containers should contain the
substance to be examined and to all containers but one increasing, known quantities of the
substances to be measured are added. This is to generate a calibration curve based on the
standard addition and thus at least three concentration levels of standard addition should be
used. The standard addition should be prepared with standard additions corresponding to,
for example, 50, 100 and 150% of the limit of the substance to be measured.
Thus solvent to be used for dissolving the compound should be prepared containing the
relevant organic solvent in different concentrations.
In the present case it could be the following preparations that should be analyzed:
Test solution: 0.50 g omeprazole dissolved in 4.0 ml N,N-dimethylacetamide.
Standard addition 1: 0.50 g omeprazole dissolved in 4.0 ml N,N-dimethylacetamide
containing 6.25 mg/ml methylene chloride and 3.125 mg/ml chloroform.
Standard addition 2: 0.50 g/ml omeprazole dissolved in 4.0 ml N,N-dimethylacetamide
containing 12.5 mg/ml methylene chloride and 6.25 mg/ml chloroform.
Standard addition 3: 0.50 g omeprazole dissolved in 4.0 ml N,N-dimethylacetamide
containing 18.75 mg/ml methylene chloride and 9.375 mg/ml chloroform.
374
Introduction to Pharmaceutical Chemical Analysis
The test solution and the standard addition solutions are analyzed by GC. Each solution
should be injected at least three times and the mean of the peak areas should be used for
further calculations. The following results are obtained:
Peak areas (mean) obtained from the chromatograms:
Sample
Test solution
Standard addition level 1
Standard addition level 2
Standard addition level 3
Methylene chloride
Chloroform
3666
13 520
23 356
36 451
4212
12 312
20 429
31 663
The peak areas in the test solution originate from the unknown amounts of methylene
chloride and chloroform in the sample. The peak areas from the standard addition
solutions originate from the unknown amounts of chloroform and methylene chloride in
the sample plus the added amounts of the same solvents.
Calculation of chloroform in sample:
Standard addition level 1 contains 4 3.125 mg/ml ¼ 12.5 mg of chloroform added
which corresponds to a peak area of 12 312 – 4212 ¼ 8100. Standard addition level 2
contains 4 6.25 mg/ml ¼ 25 mg of chloroform added which corresponds to a peak area
of 20 429 – 4212 ¼ 16 217. Standard addition level 3 contains 4 9.375 mg/ml ¼ 37.5 mg
of chloroform added which corresponds to a peak area of 31 663 – 4212 ¼ 27 451. This
means that an area of (8100 þ 16 217 þ 27 451)/3 ¼ 17 256 corresponds to an amount
of chloroform of (12.5 mg þ 25.0 mg þ 37.5 mg)/3 ¼ 25.0 mg. In the test sample solution
a peak area of chloroform was measured to be 4212, which corresponds to (4212/
17 256) 25 mg ¼ 6.1 mg chloroform. This amount was present in 0.5 g of raw material
which then corresponds to 6.1 mg/0.5 g ¼ 12.2 ppm in the raw material.
Similarly, the concentration of methylene chloride is calculated respectively to be
17.6 ppm. Both results are below the maximum limit.
Comments to the procedure:
Nitrogen is used as the carrier gas because it is inexpensive and allows separation of the
substances within an acceptable time. The column used is a capillary column with a
typical length and inner diameter. The stationary phase is semi-polar. This provides good
separation of substances with some difference in boiling point (methylene chloride
39.8 C, chloroform 61.2 C). The film thickness of stationary phase is relatively large to
give the volatile components sufficient retention. The analysis is likely to be performed
isothermally (constant temperature in the column during the chromatographic process)
because only two substances with low boiling points are to be separated. The column
temperature is not stated in the procedure, but the volatility of the two solvents requires a
column temperature to be in the range 40–60 C. A FID detector is used because this
gives a sufficient response for the two substances and because it is the most common
detector available. Samples must be equilibrated in the sample vials for 1 h at 80 C to
ensure that significant amounts of the two residual solvents has evaporated from the
sample solution and into the gas phase for analysis and has established an equilibrium.
Chemical Analysis of Drug Substances
375
added a solution of silver nitrate. If traces of chloride is present in the raw material a
precipitate of silver chloride is formed and the solution becomes turbid:
þ
ClðaqÞ
þ AgðaqÞ
! AgClðsÞ # ðprecipitateÞ
ð21:4Þ
Similarly, detection of trace amounts of sulfates in the raw material can be carried out by
addition of barium chloride to a solution of the raw material. If sulfate is present the addition
of barium ions to the sample solution will result in precipitation of barium sulfate and the
solution becomes turbid:
þ
SO4 þ Ba2ðaqÞ
! BaSO4 ðsÞ # ðprecipitateÞ
ð21:5Þ
Both in the test for chloride and in the test for sulfate an evaluation of the turbidity
obtained after addition of the precipitation reagent is needed. This is performed by
comparison of the test solution against a dilute standard solution containing sodium
chloride (for chloride) or potassium sulfate (for sulfate) prepared in a similar way as
the sample solutions. The sample and comparison sample are observed against a black
background, and if the sample is less turbid than or equal to the comparison sample, the test
is passed. An example of a monograph where tests for chlorides and sulfates are included is
shown in Box 21.19.
Box 21.19 Test for chlorides and sulfates in furosemide according to Ph.
Eur. 7th edn (2011)
Chlorides (2.4.4): maximum 200 ppm.
To 0.5 g add a mixture of 0.2 ml of nitric acid R and 30 ml of water R and shake for 5 min.
Allow to stand for 15 min and filter.
Sulfates (2.4.13): maximum 300 ppm.
To 1.0 g add a mixture of 0.2 ml of acetic acid R and 30 ml of distilled water R and shake
for 5 min. Allow to stand for 15 min and filter.
Discussion of the procedures:
The figures in brackets refer to the general monographs for test of chlorides and sulfates,
respectively. This monographs therefore have to be consulted in order to be able to
perform the test.
Chlorides: furosemide is practically insoluble in water and dilute acid and the chloride
impurities therefore have to be extracted and the mixture filtered to obtain a clear solution.
15 ml of this filtrate is added 1 ml of dilute nitric acid R (125 g/l), and the total solution
is then poured into a test tube containing 1 ml of silver nitrate solution R2 (17 g/l).
A standard is prepared in the same manner using 10 ml of chloride standard solution
(5 ppm Cl) R and 5 ml of water. This mixture is poured into a test tube containing 1 ml of
376
Introduction to Pharmaceutical Chemical Analysis
silver nitrate solution R2 (17 g/l). Both tubes are observed vertically against a black
background (see Box 21.17). After standing for 5 min protected from light, any opalescence in the sample are not more intense than in the standard.
Sulfates: 1 g of furosemide is extracted with 30 ml of distilled water added 0.2 ml of
concentrated acetic acid. 15 ml of the filtrate is removed and added to 0.5 ml of acetic
acid and 2.5 ml of barium chloride solution (250 g/l) prepared in the following way: 3 ml
of a 250 g/l solution of barium chloride R is added to 4.5 ml of sulfate standard solution
(10 ppm SO4) R1. A standard is prepared in the same manner using 15 ml of sulfate
standard solution (10 ppm SO4) R instead of the solution to be examined. The opalescence in the solutions is compared after 5 min.
To obtain a repeatable precipitation of barium sulfate a small amount of sulfate has to be
present in advance in both solutions.
Calculation of limits of content:
In the test for chloride a maximum limit of 200 ppm is set. This means that chloride ions
in amounts up to 200 ppm are allowed in the raw material. The comparison standard
correspond to 200 ppm according to the following calculation:
The content of chloride ions in the standard is:
½10 ml=ð10 ml þ 5 ml þ 1 mlÞ 5 ppm ¼ 3:125 ppm ¼ 3:125 mg=ml
Thus not more than 3.125 mg chloride ions per ml is allowed in the test solution:
3.125 mg/ml in (15 ml þ 1 ml) corresponds to16 ml 3.125 mg/ml ¼ 50 mg
Test solution contains half of the amount of the raw material that was weighed (15 ml of
30 ml were used for the test solution). This means that 50 mg is allowed in 0.25 g of the
raw material which corresponds to 50 mg/0.25 g ¼ 200 ppm.
21.5.1.8 Cationic Impurities
In addition to anionic impurities most monographs also include one or more purity tests to
discover any trace of foreign cations. The content of cationic impurities has to be limited
because they can have toxic effects (heavy metals) or because they can have catalytic
effects, which can result in a faster degradation of the drug substance. The presence of
cationic impurities is determined by two different tests. One method is based on the control
of heavy metals and this test is called “Heavy metals.” This test is relatively specific for
heavy metals like Pb, Cu, Fe and Hg. The second method is called “Sulfated ash.” This
method is not specific to the cations, but controls the overall level of inorganic impurities in
the raw material. The test for heavy metals is based on the fact that heavy metals form
colored sulfides that are practically insoluble. In Ph.Eur. there are eight different methods
for purity testing of heavy metals, but focus here is only on the two simplest methods, called
methods A and B. In method A the raw material is dissolved in an aqueous medium, and a
buffer is added to stabilize pH at 3.5. If the pH is too low the sulfides will not precipitate. To
this solution thioacetamide reagent R [0.2 ml of thioacetamide solution (40 g/l) þ 1 ml of a
Chemical Analysis of Drug Substances
377
mixture consisting of 5 ml of water, 15 ml of 1 M NaOH and 20 ml of glycerol] is added.
The thioacetamide will form sulfide ions in the solution according to the following
reactions:
CH3 CSNH2 þ H2 O ! CH3 CONH2 þ H2 S ð1Þ
ð21:6Þ
H2 S ! H þ þ HS ! 2H þ þ S2 ð2Þ
ð21:7Þ
If the raw material contains trace amounts of heavy metals, these combine with the sulfide
ions forming a precipitate of metal sulfides. The reaction with lead is chosen as an example:
þ
Pb2ðaqÞ
þ S2ðaqÞ
! PbSðsÞ ð#Þ
ð21:8Þ
The precipitate provides the solution with a brown color which is compared to a standard
lead solution treated in a similar way. The comparison test is usually made by adding
thioacetamide reagent to a very dilute solution of lead nitrate. The test for heavy metals
include all metals that form insoluble sulfides, including lead, copper, silver, mercury,
cadmium, bismuth, rutenium, gold, platinum, palladium, vanadium, arsenic, antimony, tin
and molybdenum.
In test method B for heavy metals, the same procedure similar to method A is used, except
that the raw material is dissolved in a mixture of an organic solvent (miscible with water) and
water. This method is designed for raw materials that are not easily soluble in water. An
example of a purity test for heavy metals using method B is shown in Box 21.20.
Box 21.20 Test for heavy metals in ibuprofen according to Ph.Eur. 7th
edn (2011)
Heavy metals (2.4.8): maximum 10 ppm.
12 ml of solution S complies with test B. Prepare the reference solution using lead
standard solution (1 ppm Pb) obtained by diluting lead standard solution (100 ppm Pb) R
with methanol R.
Solution S: Dissolve 2.0 g in methanol R and dilute to 20 ml with the same solvent.
Discussion of the procedure:
To 12 ml of solution S, 2 ml of ammoniumacetate-HCl buffer pH 3.5 and 1.2 ml of
thioacetamide reagent are added.
A reference solution is prepared from 10 ml of lead standard solution (1 ppm Pb), 2 ml of
solution S, 2 ml of ammoniumacetate-HCl buffer pH 3.5 and 1.2 ml of thioacetamide
reagent.
The 1 ppm lead standard solution is prepared by dilution of a 100 ppm lead solution using
methanol for dilution because this solvent was used for preparation of solution S.
A blank solution is prepared using 10 ml of methanol, 2 ml of solution S, 2 ml of
ammoniumacetate-HCl buffer pH 3.5 and 1.2 ml of thioacetamide reagent.
378
Introduction to Pharmaceutical Chemical Analysis
After mixing all three solutions are compared after 2 min. Assessment is made against a
white background.
The reference solution should show a slight brown color compared to the blank solution.
Any brown color in the test solution is not more intense than that in the reference
solution.
If the result is difficult to judge the solutions may be filtered through a 0.45 mm membrane
filter, and the spots can then be compared on the membrane filters.
Calculation of limits of content:
The test for heavy metals in ibuprofen is set at 10 ppm. This limit can also be calculated in
the following way.
In the comparison sample, the total content of heavy metals (Pb) is 10 mg when
considering the 10 ml of lead standard solution (1 ppm Pb) used. In the test solution 12 ml
of solution S is used, but as 2 ml of this solution is also added to the comparison solution
only 10 ml is considered in the calculations. Thus 10 mg Pb is allowed in the solution
S. Solution S is prepared by extraction of 2.0 g of ibuprofen into 20 ml (¼ 1.0 g/10 ml).
The limit is 10 mg Pb in 10 ml solution S. In the sample solution this corresponds to
10 mg/g ¼ 10 ppm.
As previously mentioned, the test for heavy metals is relatively specific, and the test is
present in many monographs. However, modern synthesis techniques and equipment have
made this test relatively redundant as the purity of APIs often is high and the test is therefore
most often negative. In contrast a lot of new catalysts are now used in modern synthesis and
impurities from these catalysts may not be detected by the heavy metals test. Therefore new
ways to detect traces of elements in raw materials are now considered and among these are
techniques like ICP-OES and ICP-MS.
A test for the determination of sulfated ash is included in many Ph.Eur. monographs. This
test gives a general picture of the level of inorganic impurities in a raw material. In some
cases this test is in addition to the test for heavy metals and foreign inorganic anions, while in
other monographs only the test for sulfated ash is prescribed. The test is performed by
placing a small amount of raw material in a crucible. The raw material is moistened with
sulfuric acid and the crucible is heated. During this process, all organic material is
combusted and disappears primarily as CO2 and H2O, while inorganic cations remain in
the crucible as sulfates (sulfated ash). The exact weight of both the crucible and the raw
material is determined before and after the ashing procedure. The weight of the sulfated ash
should generally not exceed 0.1% (w/w) of the raw material. An example of the determination of sulfated ash is given in Box 21.21.
21.5.1.9 Loss on Drying
In addition to sulfated ash most of the monographs contain a test either for “Loss on drying”
or for “Water.” In the test for loss on drying an exact amount of raw material is dried either in
an oven at 105 C (or other temperature specified in the monograph), in a desiccator
containing P2O5 or under vacuum at a given temperature. After drying the sample is weighed
Chemical Analysis of Drug Substances
379
Box 21.21 Determination of sulfated ash in paracetamol according to
Ph.Eur. 7th edn (2011)
Sulfated ash (2.4.14): maximum 0.1%, determined on 1.0 g.
Paragraph 2.4.14 gives the procedure for determination of sulfated ash:
A crucible of either platinum, porcelain or quartz is heated to 600 C for 30 min and then
cooled in a desiccator containing silica gel (silica gel absorbs any moisture). The crucible
is weighed accurately on an analytical balance. 1.0 g of raw material (paracetamol) is
placed in the crucible, and the exact weight of the crucible and raw materials is
determined on the analytical balance. The weight of the raw material is determined
as the difference between the two weighings. Then the raw material is moistened with
1 ml of 1 M sulfuric acid, and the crucible is heated until the raw material is carbonized.
The crucible is cooled, and the residue is moistened with 1 M sulfuric acid (usually 1 ml).
The crucible is then heated again until white fumes are no longer evolved from the
sample, and then the crucible is ignited at 600 C 50 C until the residue is completely
incinerated. Finally, the crucible is allowed to cool in a desiccator containing silica gel.
Weigh the crucible and calculated the percentage of residue (sulfated ash).
Calculation example:
1.0660 g of raw material is transferred to a crucible that weighs 23.5664 g. After the
ashing procedure the weight of the crucible is found to be 23.5691 g. The sulfated ash
determined corresponds to:
Weight of sulfate ash ¼ 23:5671g 23:5664g ¼ 0:0007g
% of raw material ¼ ð0:0007g=1:0660gÞ 100% ¼ 0:07%
Thus the paracetamol raw material is within specification.
again and the loss can be calculated. Volatile compounds such as water and residual solvents
evaporate during drying, and the weight loss determined is equal to the sum of these
components. Thus using the test “Loss on drying” the amount of volatile components is
determined, and it is a very simple way to get an overview of the overall level of water and
solvents in a compound. If it is necessary to have a more specific test for residual solvents the
gas chromatographic procedure preciously described (Section 21.5.1.6) should be used.
When a more specific test for water is needed the titration procedure described below
is used.
An example of the determination of loss on drying is shown in Box 21.22.
21.5.1.10 Determination of Water
In many cases loss on drying is sufficient to control the water content. Some raw materials
are particularly sensitive to traces of water because this can accelerate the decomposition of
the raw material. Other substances are hydrates in which a well defined content of water
380
Introduction to Pharmaceutical Chemical Analysis
Box 21.22 Determination of loss on drying in paracetamol according to
Ph.Eur. 7th edn (2011)
Loss on drying (2.2.32): maximum 0.5%, determined on 1.000 g by drying in an oven at
105 C.
Discussion of the procedure:
The full information on how to perform the test is given in paragraph (2.2.32).
A glass weighing bottle is dried to constant weight in an oven at 105 C and weighed. The
prescribed quantity of the substance to be examined is then placed in the weighing bottle
and weighed. The weight of the sample is the difference between the two weighings. Dry
the substance to constant mass or for the prescribed time. In the case for paracetamol
drying to constant mass is required as no other information is given in the monograph.
Each time before weighing the weighing bottle has to be placed in a desiccators to cool to
room temperature. The weight is considered to be constant when the difference between
two weightings is less than 0.0005 g. When the drying temperature is indicated by a
single value the drying is carried out at the prescribed temperature 2 C. In this case it
is 105 2 C.
Calculation example:
1.1023 g of raw material is transferred to a weighing glass that weighs 18.2361 g. After
drying to constant weight the glass weigh 19.3361 g. The loss on drying is calculated
to be:
Weight ¼ 18:2361g þ 1:1023g 19:3361g ¼ 0:0023g
% of raw material ¼ ð0:0023g=1:1023gÞ 100% ¼ 0:21%
Thus the paracetamol raw material is within specification.
should be present. For such compounds a separate and more specific, quantitative test for
water can be prescribed. In the monographs this test is called “Water,” and it is based on the
Karl–Fischer method. A small amount of the raw materials typically dissolved in anhydrous
methanol is gradually added Karl–Fischer reagent from a burette. The reagent consists of a
solvent such as pyridine, iodine in a known concentration, sulfur dioxide in excess relative to
the amount of water in the analyte and methanol. The following reaction takes place
between the water and the Karl–Fischer reagent:
H2 O þ C5 H5 N I2 þ C5 H5 N SO2 þ C5 H5 N ! 2C5 H5 N HI þ C5 H5 N SO3 ð1Þ ð21:9Þ
C5 H5 N SO3 þ CH3 OH ! C5 H5 NðHÞSO4 CH3 ð2Þ
ð21:10Þ
In reaction (1) oxidation of SO2 by I2 to SO3 consumes water that is present in the raw
material and I2 is reduced to HI. There is a large excess of pyridine, and SO2, I2, SO3 and HI
Chemical Analysis of Drug Substances
381
will be present as complexes with pyridine. These are marked with . in the reaction schemes.
Reaction (2), which occurs when an excess of methanol is added, is important because the
sulfur trioxide complex with pyridine also consume water according to the following
reaction:
C5 H5 N SO3 þ H2 O ! C5 H5 NHSO4 H ð3Þ
ð21:11Þ
Reaction (3) is not specific for water, and is therefore undesirable. It is prevented by
having a large excess of methanol in the solution to be titrated. During the titration a color
change of the solution is observed. When all the water in the raw material has reacted with
the Karl–Fischer reagent a further addition of the reagent results in the presence of unreacted
reagent (brown color) in the sample solution. This gives a color change from yellow to
brown. The color change show how much volume of reagent is needed to “neutralize” the
water in the sample and the amount of water can be calculated. Alternatively, an
electrochemical detection method can be used for detection of the endpoint, where all
water in the raw material has reacted. At the endpoint a significant change in current passing
through the solution can be measured. Karl–Fischer titration is therefore not a limit test but a
quantitative method that accurately and specifically determines the water content of a raw
material. Box 21.23 shows an example of the test “Water.”
Box 21.23 Determination of water in ephedrine according to Ph.Eur. 7th
edn (2011)
Water (05/02/1912). Not more than 0.5%, determined on 2.000 g by the semi-micro
determination of water.
Main features of the procedure:
To 20 ml of anhydrous methanol Karl–Fischer reagent is gradually added from the
burette (titrated). Two micro electrodes are placed in the solution and a small electrical
voltage is imposed across them. Current passing through the solution is measured
continuously while the Karl–Fischer reagent is added. Karl–Fischer reagent is added
until all water has reacted (endpoint). When the endpoint is reached, no more water
remains in the methanol solution. 2.000 g of raw material is then dissolved in methanol,
and Karl–Fischer reagent is again added until it again reaches an endpoint. It is read how
much reagent was used, and from this it is calculated how much water was present in the
raw material. Water content must not exceed 0.5% (w/w) of the raw material.
Calculation example:
2.1016 g of raw material is dissolved in 20 ml of methanol, and titrated with 1.25 ml of
Karl–Fischer reagent which is 0.1562 M with respect to I2.
Consumed quantityI2 ¼ ð1:25 ml=1000 mlÞ 0:1562 M ¼ 1:95 104 mol
382
Introduction to Pharmaceutical Chemical Analysis
Water in raw materials ¼ 1:95 104 mol ¼ 1:95 104 mol 18:02g=mol
¼ 3:52 103 g
Water content ð%Þ ¼ ð3:52 103 g=2:1016gÞ 100% ¼ 0:17% ðw=wÞ
The water content is below the permitted level and the test is approved.
21.5.2
Purity Tests for Raw Materials of the Type of Mixtures of Organic
Compounds
Identification and purity testing are also needed for ingredients that are mixtures of organic
compounds and this involves a number of different test procedures of which some are
different from those used for well defined chemicals. This section focuses on the most
important tests, some of which are used for both identification and for purity testing and are
therefore discussed together. These tests are summarized in Table 21.11. In addition,
microscopy, viscosity and density measurements can also be used.
Table 21.11 Overview of important tests for identification and purity testing of raw materials of
the type of mixtures of organic compounds
Test method
What is measured
by the method
Purpose of test
method
Specificity
Appearance of solution
Impurities of low
solubility and
colored impurities
Physical constants
Purity testing
Low
Identification and
purity testing
Purity testing
Low
Low
Low
Freezing, melting or
boiling point
Acidity/alkalinity or pH
Oxidizing substances
Acid value
Acidic or basic
impurities
Oxidants
Free acid
Low
Peroxide value
Saponification value
Number of groups
to be acetylated
Unsaturated
compounds
Peroxides
Esters
Unsaponifiable matter
Specified impurities
Named impurities
Purity testing
Identification and
purity testing
Identification and
purity testing
Identification and
purity testing
Purity testing
Identification and
purity testing
Purity testing
Purity testing
Composition of
fatty acids
Identification and
purity testing
Heavy metals
Water
Volatile
substances
Inorganic matter
Purity testing
Purity testing
Purity testing
Low
Specific
Low
Purity testing
Low
Hydroxyl value
Iodine value
Composition of fatty
acids (gas
chromatography)
Heavy metals
Water
Loss on drying
Sulfated ash
Low
Low
Low
Low
Low
Relatively
specific
Highly specific
Chemical Analysis of Drug Substances
383
The tests for appearance of solution, melting point and acidity/alkalinity/pH are the same
as described in Section 21.5.1.3 and are not further discussed here. The principle of the tests
for the freezing point and boiling point do not differ significantly from the test for the
melting point, and are not discussed further.
21.5.2.1 Oxidizing Substances
Using the test for oxidizing substances it is verified that the raw material does not contain
significant amounts of compounds that can oxidize potassium iodide according to the
following reaction:
2 I þ oxidizing impurity ! I2 þ reduced impurity ð1Þ
ð21:12Þ
An excess of potassium iodide is added to a known quantity of the raw material. A part of
the iodide ions will react with the oxidizing impurity and form iodine, I2. The iodine can then
be determined by titration with sodium thiosulfate and starch as an indicator according to the
following reaction:
2
I2 þ 2S2 O2
3 ! 2I þ S4 O6 ð2Þ
ð21:13Þ
Although the test is used for purity testing it is performed as a quantitative measurement. The test is important for some substances because the presence of oxidizing
compounds can reduce the stability of the ingredients in a finished product. For some
substances there is also a test for reducing substances. Several different principles are used
for this test, and it is used on a relatively small number of substances, and is therefore not
discussed further.
21.5.2.2 Acid Value
The acid value is determined on a number of substances. The purpose of the test is to
determine the total amount of free acids in the raw material. The acid value is defined as the
amount of potassium hydroxide that must be added to neutralize the free acids in 1 g of the
raw material. The procedure is to weigh m g of the raw material and dissolve it in a mixture of
equal amounts of ethanol (96%) and light petroleum R3 (boiling range 100–120 C). This
solution is titrated with n ml of 0.1 M KOH or 0.1 M NaOH to the color changes from
colorless to pink with phenolphthalein as indicator. The acid value (IA) is calculated in the
following way:
IA ¼ mg KOH required to neutralize the free acids present in 1 g of the substance.
IA ¼
5:610 n
m
ð21:14Þ
Where m is the mass (g) of substance and n is the volume (ml) of 0.1 M KOH or 0.1 M
NaOH. One mole of KOH is 56.1 g/mol. Observe that the acid value is expressed as x mg
KOH even if 0.1 M NaOH is used for titration.
The acid value is often determined in fatty oils and synthetic polymers to ensure that the
raw material does not contain too high levels of unreacted free acids. For fatty oils a high
acid value indicates that there are a lot of free fatty acids present. An upper limit on the acid
value is therefore given in each monograph.
384
Introduction to Pharmaceutical Chemical Analysis
21.5.2.3 Hydroxyl Value
In addition to the acid value a test that determines the hydroxyl value can be found in many
monographs. In this test a value correlating to the amount of acetic acid consumed by
acetylation is determined. This is particularly relevant for a variety of synthetic polymers,
where the number of free OH groups depends on the degree of polymerization. One can thus
verify if the degree of polymerization is appropriate by determination of the hydroxyl value.
The hydroxyl value (IOH) is defined as the number of milligrams of potassium hydroxide
required to neutralize the acid liberated by acylation of 1 g of the substance. The acylation is
performed using acetic acid anhydride. During this reaction the following happens:
R CH2 OH þ ðCH3 COÞ2 O ! R CH2 OCOCH3 þ CH3 COOH
ð21:15Þ
For each mole of OH groups in the raw material 1 mol of acetic acid is produced in the
reaction. The exact amount of acetic acid is determined by titrating with 0.5 M alcoholic
potassium hydroxide solution (n1 ml) until a color change with phenolphthalein from
colorless to pink is observed. A blank test is carried out under the same conditions (n2 ml of
0.5 M alcoholic potassium hydroxide solution is consumed). IA is the acid value determined:
IOH ¼
28:05 ðn2 n1 Þ
þ IA
m
ð21:16Þ
Any free acid is of course also neutralized by the potassium hydroxide and is included in
the n1 volume which is subtracted from the blank. Therefore the acid value has to be added in
the calculation.
The requirement to the hydroxyl value is typically given as an interval in the monographs.
21.5.2.4 Iodine Value
The iodine value (II) is another test performed for a variety of substances like synthetic
polymers and hydrocarbons. The iodine value expresses the degree of unsaturation in 100 g
of the substance, that is, the extent to which various compounds in the raw material contain
carbon–carbon double bonds. The iodine value is defined as the number of grams of iodine
that can be consumed by (substituted into) 100 g of raw material. An accurately weighed
quantity of raw material (m g) is dissolved in chloroform, and a known quantity of an iodine
bromine solution (IBr) is added. The following reaction occurs between the double bonds in
the raw material and IBr (IBr is added to the C–C double bonds):
C ¼ C þ IBr ! CI CBr
ð21:17Þ
The surplus of IBr reacts with potassium iodide according to the following reaction:
IBr þ I ! I2 þ Br
ð21:18Þ
The iodine formed by this reaction can be titrated with 0.1 M sodium thiosulfate
according to the following reaction (n1 ml of thiosulfate solution are consumed):
I2 þ 2S2 O23 ! 2I þ S4 O2
6
ð21:19Þ
A blank test is carried out under the same conditions (n2 ml of thiosulfate solution
consumed).
Chemical Analysis of Drug Substances
385
The amount of IBr that has reacted with the raw material can now be calculated using the
difference between the volume of the blank and the volume corresponding to the amount of
IBr that did not react:
II ¼
1:269 ðn2 n1 Þ
m
ð21:20Þ
The value gives an indication of the amount of double bonds contained in the raw
material. The iodine value is often a test included in monographs containing oils and fats as
well as in monographs like oleic acid and stearic acid. In the monographs the requirement to
the iodine value is typically a maximum number not to be exceeded.
21.5.2.5 Peroxide Value
Peroxides are undesirable in large quantities in raw materials because they behave as
oxidants and therefore can affect the stability of the raw material itself or other
components it may be mixed with in a drug formulation. Peroxide value is a test used
to determine the amount of peroxides in the ingredients. The principle of the test is that to
a known quantity of raw material a surplus of potassium iodide is added for the following
reaction to take place:
ROOH þ 2I þ 2H þ ! ROH þ I2 þ H2 O
ð21:22Þ
The amount of iodine formed is subsequently determined by titrating with 0.01 M sodium
thiosulfate according to the following reaction:
I2 þ 2S2 O23 ! 2I þ S4 O26 ð21:23Þ
Starch is used as a color indicator to determine the titration endpoint. The peroxide value
is expressed as the number of milliequivalents of active oxygen in 1000 g of the substance,
and the number is usually located within the range 1–100. The requirement for the peroxide
value vary from substance to substance and is set as an upper limit and the test is often
present in monographs for synthetic polymers and fatty oils.
21.5.2.6 Saponification Value
Determination of saponification value is present as a test in some raw materials and
especially among the synthetic polymers and fatty oils. The saponification value (IS) is
defined as the number of milligrams of potassium hydroxide needed to neutralize free acids
and to saponify (hydrolyze) esters present in 1 g of the raw material. The procedure is to
dissolve a known quantity (m g) of the raw material in 25.0 ml of alcoholic potassium
hydroxide. The potassium hydroxide has to be in excess. The mixture is refluxed for 30 min
unless otherwise prescribed. The following reactions occur:
R1 COOH þ OH ! R1 COO þ H2 O
ðneutralization of free acidsÞ
R2 COOR3 þ OH ! R2 COO þ R3 OH
ðhydrolysis of estersÞ
ð21:24Þ
ð21:25Þ
The amount of potassium hydroxide is added in excess, and the remaining potassium
hydroxide is titrated with 0.5 M hydrochloric acid (n1 ml) using phenolphthalein as
386
Introduction to Pharmaceutical Chemical Analysis
indicator. A blank test is carried out under the same conditions (n2 ml of 0.5 M hydrochloric
acid consumed).
From these titrations, the amount of reacted potassium hydroxide is calculated, and the
saponification value can be determined:
IS ¼
28:05ðn2 n1 Þ
m
ð21:26Þ
The requirement for the saponification value is usually given as an interval, and if the
saponification value falls outside the interval, the raw material is out of specification.
21.5.2.7 Unsaponifiable Matter
Determination of unsaponifiable matter is a test that is used to characterize fatty oils that are
composed of triglycerides. The procedure involve saponification of a known amount (m g)
of the raw material in the first place:
triglyceride þ KOH ! glycerol þ potassium salts of fatty acids
ð21:27Þ
After the saponification the reaction mixture is diluted with water and extracted with
peroxide-free diethyl ether in a separating funnel. Glycerol and the potassium salts of the fatty
acids will remain in the water phase while lipophilic constituents that could not be hydrolyzed
will be extracted to the ether phase. The ether phase is purified by extraction with potassium
hydroxide solution and water to neutrality and then evaporated to dryness. The weight of the
residue (a g) is determined and the content of unsaponifiable matter is calculated:
Unsaponifiable matter ¼
100 a
per cent
m
ð21:28Þ
Lipophilic organic compounds (e.g., hydrocarbons), which was not originally on the ester
form, and that has not been saponified, however, will be extracted into diethyl ether, and the
amount is typically not more than 1–2% (w/w) of the raw material.
21.5.2.8 Other Tests
In raw materials of the type of fatty oils and essential oils, which are based on mixtures of
different triglycerides or terpenes, respectively, it is customary to control the composition
using GC. For fatty oils containing triglycerides, the triglycerides are hydrolyzed, and the
fatty acids released are transformed into methylesters. The methylesters are analyzed by GC
in order to give the composition of fatty acids. The main components are determined
quantitatively by the normalization procedure (see Chapter 20). When analyzing essential
oils a small amount of the oil is injected into the GC system, and the components are
separated. Also a standard solution containing known amounts of the main components is
injected to the GC. The main components of the oil can in this way be identified based on
retention time, and the single component can also be quantified. The requirements for the
main components are given in the respective monograph as ranges or limits of the amounts
allowed for each of the main components. An example of an essential oil is shown in
Box 21.24.
There are many examples of specific impurities in the monographs. These vary widely in
nature, and the tests used are very different depending on the impurity to be controlled.
These are not discussed further in this text.
Chemical Analysis of Drug Substances
387
Box 21.24 Identification of peppermint oil by GC according to Ph.Eur.
7th edn (2011)
Main features of the procedure:
0.20 g of peppermint oil is diluted to 10.0 ml with hexane and injected into the GC
system. A reference solution is prepared by dissolving 10 ml of limonene, 20 ml of
cineole, 40 ml of menthone, 10 ml of menthofuran, 10 ml of isomenthone, 40 ml of menthyl
acetate, 20 ml of isopulegone, 60 mg of menthol, 20 ml of pulegone, 10 ml of piperitone
and 10 ml of carvone to 10.0 ml with hexane. 1 ml of the test and reference solutions are
injected separately into a GC system consisting of the following main components:
Carrier gas: 1.5 ml/min helium
Column: 60 m long, 0.25 mm inner diameter with a stationary phase of macrogol 20 000.
Column temperature program: 60 C for 10 min, then increase the temperature by 2 C/min
to 180 C. This temperature is kept for 5 min before the oven cools down again to 60 C.
Detection: Flame ionization detector, at 220 C
Injection: Split injection 1 : 50 (2% of the sample is transferred to the column while the
remainder is vented out) at 200 C
Requirements for positive identification:
Retention of the characteristic (main) peaks in the chromatogram of peppermint oil are
similar in retention time to those of peaks in the chromatogram obtained with the reference
solution. The figure below shows an example of a chromatogram of peppermint oil.
Peaks 1 ¼ limonene, 2 ¼ cineol, 3 ¼ menthone, 4 ¼ menthofuran, 5 ¼ isomenthone,
6 ¼ menthyl acetate, 7 ¼ Menthol, 8 ¼ pulegon
The identity test refers to the test for Chromatographic profile given under “Tests”.
This means that the same method is used for identification and for the chromatographic
profile. The full description of the method is given under Chromatographic profile in the
Ph.Eur. monograph on peppermint oil.
388
Introduction to Pharmaceutical Chemical Analysis
Comments on the procedure:
Helium is used as a carrier gas because it provides rapid analysis compared to nitrogen,
and because it is more safe to use than hydrogen. The split injection is used because a
sample volume of 1 ml is too large for a column with an internal diameter of only
0.25 mm. With a split ratio of 1 : 50, only 2% of the sample is introduced into the
column, thus ensuring that the column is not overloaded with subsequent loss of
chromatographic resolution. The temperature in the injector set at 200 C to ensure that
all test components are quickly transferred to the gas phase and introduced into the
column. The column used is a relatively long capillary column, and the length is
selected to provide high efficiency of the many analytes present in the sample. The
stationary phase is medium polar and separate the analytes according to both volatility
and polarity. The GC analysis is performed using temperature programming, which
means that the temperature is changed during chromatography. Initially the temperature is kept at 60 C for 10 min. The most volatile components will eluted through the
column, while less volatile analytes have a much higher affinity to the stationary phase
at the entrance of the column. Then the temperature is increased gradually. The
temperature increase is low (2 C/min), which provides very good resolution in the GC
system. At the end the temperature is kept for 5 min at 180 C to ensure that all also less
volatile sample components are eluted from the column before the system cools down
to its initial temperature to be ready for the next analysis. Flame ionization detection is
used as this detector has a large linear range and responds to all components. In this
case it is not necessary with a more selective detector. The temperature of the detector
is kept at 220 C to ensure that no sample components condense in the detector.
As with HPLC, GC is a highly efficient separation method, and identification based on
GC is very specific.
Heavy metals, water, loss of drying and sulfated ash listed in Table 21.11 are discussed in
Section 21.5.1, and are not treated further in this section. Total ash is basically equivalent to
the test for sulfated ash, except that sulfuric acid is not added during the ashing procedure.
This means that the ash that is left consists of various inorganic salts and oxides.
21.5.3
Identification of the Raw Materials of the Type of Mixtures of Organic
Compounds
In addition to ingredients that are well defined chemical compounds the identification of raw
materials which are mixtures of organic compounds shall also shortly be discussed. This
group of substances includes a wide variety of different products which can roughly be
divided into polysaccharides, synthetic polymers, fatty oils and hydrocarbons. With few
exceptions, all these substances are used as excipients in pharmaceutical preparations. A
relatively large number of identification procedures are used for this group of ingredients.
Only those tests used frequently are described here, and the tests are summarized in
Table 21.11. For mixtures of organic compounds, no distinction is normally made between
first and second identification, and for some compounds no identification procedures is
given at all.
Chemical Analysis of Drug Substances
389
Section 21.4.2 discusses the use of thin-layer chromatography (TLC), optical rotation and
melting point. Color reactions and different kinds of precipitation reactions are often used in
mixtures of organic compounds, butthe principles behind thesevary greatly from substance to
substance. The acid value, ester value, hydroxyl value, iodine value and peroxide value are
simple, chemical methods used for characterization of a given chemical functionality of a
substance; and these tests, together with hydroxyl value and saponification value, are
discussed in more detail in Section 25.5.2. All the tests are used for identification as well
as limit tests and therefore each test is often described under “Tests,” and under
“Identification” there is then a cross-reference to the “Tests” section. The physical test for
viscosity, the evaluation of appearance and microscopy are not further discussed in this text.
Gas chromatography (GC) is used extensively for identification in connection with the
fatty oils essential oils. GC is presented in Chapter 14. The identification of fatty oils by GC
includes a chemical derivatization where the fatty acids in the triglycerides are liberated
under the formation of methylesters of the fatty acids. The methyl esters formed are then
analyzed by GC using a standard mixture of fatty acid methyl esters as reference standard. In
the identification of essential oils the oil itself is injected into the GC without any sample
preparation or after a simple dilution with hexane. The chromatogram is compared to a
standard mixture of aroma compounds relevant for the given essential oil. It is common to
first inject a solution of a standard that contains one or more of the main components of the
oil and then inject a volume of the oil. For a positive identification the characteristic (main)
peaks in the chromatogram of the oil should be similar in retention time to peaks of the
reference solution. An example of the use of GC for the identification of an essential oil is
shown in Box 21.24.
21.6 How to Determine the Purity of Pharmaceutical Raw Materials
As previously mentioned monographs for raw materials of the type well defined chemical
compounds contain an assay for quantitative determination by which the content or purity is
determined as a percentage (w/w). Normally the requirement to the content is very strict, and
it is often in the range 99.0–101.0% as previously shown in Table 21.5. The range for the
content is, however, also dependant on the precision of the analytical chemical technique used
for the assay. Thus, titrations can normally be performed with a precision where three times
the standard deviation are within 1% and the range can therefore be limited to 99.0–101.0%.
The precision of an assay performed by the use of HPLC at several laboratories may be less
good compared to within laboratory precision and therefore the range for content is
98.0–102.0% when the assay is based on HPLC. Using UV spectrophotometry the range
is even wider: 97.0–103.0%. The lower limit of the range may sometimes be even lower to
allow for some impurities to be present. That the upper limit can be above 100% is due to
analytical uncertainty as well as the lack of specificity of titration and UV spectrophotometry.
The most important assay techniques are summarized in Table 21.12.
21.6.1
Acid–Base Titration in Aqueous Environment
Many compounds are determined quantitatively by acid–base titration in an aqueous
environment. The principle of this is discussed in Chapter 5. An accurately weighed
390
Introduction to Pharmaceutical Chemical Analysis
Table 21.12 Overview of the important techniques for quantitative determination of raw
materials of the type well defined chemical compounds
Technique
Specificity
Acid–base titration in aqueous environment
Acid–base titration in nonaqueous environment
Redox titration
HPLC
UV spectrophotometry
Low
Low
Low
High
Low
Standard range for
content (%)
99.0–101.0
99.0–101.0
99.0–101.0
98.0–102.0
97.0–103.0
quantity of the raw material is dissolved in a suitable solvent. In many cases, this solvent is
not only water, since many drugs are poorly soluble in water. Therefore the compounds are
dissolved in a mixture of water and a polar organic solvent, or in some cases in pure
organic solvent. The solvent used is typically methanol, ethanol or dimethylformamide,
all of which are miscible with water. If the compound is a basic substance it is titrated with
an aqueous solution of an acid (the titrant) of known concentration according to the
following reaction:
B þ H3 O þ ! BH þ þ H2 O
ð21:29Þ
The titrant is added gradually from a volumetric burette and when the endpoint is reached
the volume of titrant is read on the burette. On the basis of the consumption of acid the
amount of the compound in the sample solution is calculated. Based on this figure and the
exact weight of the raw material, one can calculate the purity of the compound as a
percentage (w/w).
Many APIss are basic compounds that exist as hydrochlorides. The bases are protonated
and they behave as acids in an aqueous environment. Hydrochloride salts (and hydrobromide salts) as well as other acids can be titrated with an aqueous solution of a base of
known concentration according to the following reaction:
HA þ OH ! A þ H2 O
ð21:30Þ
An example of a quantitative determination of an acidic raw material by acid–base
titration in aqueous environment is shown in Box 21.25. Similarly, an example of
quantitative determination of a hydrochloride of a basic raw material using acid–base
titration in aqueous environment is shown in Box 21.26.
In the titrations prescribed in Ph.Eur. the aim is to consume about 8 ml of titrant. This is
important in order to ensure that the volume of added titrant is not too small, if so the relative
uncertainty increases.
Some amines are often more easily dissolved in dilute acid and therefore the titration with
base becomes a little different. The basic compound will consume an equivalent amount of
protons according to the following reaction before the titration take place:
B þ H þ ! BH þ
ð21:31Þ
Sample solution is titrated with an aqueous solution of a strong base to the equivalent
point. The consumption of base is read and this titrant volume corresponds to the surplus of
Chemical Analysis of Drug Substances
391
Box 21.25 Quantitative determination of omeprazole by acid–base
titration in aqueous environment according to Ph.Eur. 7th edn (2011)
Assay: Dissolve 1.100 g in a mixture of 10 ml of water R and 40 ml of 96% ethanol R.
Titrate with 0.5 M sodium hydroxide, determining the endpoint potentiometrically
(2.2.20).
1 ml of 0.5 M sodium hydroxide corresponds to 0.1727 g of C17H19N3O3S. The
molecular mass of omeprazol is 345.4 and therefore 1 ml of 0.5 M sodium hydroxide
is equivalent to 172.7 mg of the substance.
Discussion of the procedure:
1.100 g raw material (HB) is dissolved in a mixture of 10 ml of water and 40 ml of
ethanol. This is an aqueous environment. The solution is titrated with 0.5 M sodium
hydroxide (OH). The endpoint is determined potentiometrically.
Reaction:
HB þ OH ! B þ H2 O
Example of calculation:
1.1025 g of raw material is dissolved and titrated with 6.35 ml of 0.4989 M NaOH. The
molecular weight of omeprazole is 345.4 g/mol. The purity of the raw material is
calculated as follows:
The number of moles of
0.4989 M ¼ 3.168 103 mol
base
(OH)
consumed ¼ (6.35 ml/1000 ml) The number of moles of omeprazole ¼ 3.168 103 mol
Omeprazole in g ¼ 3.168 103 mol 345.4 g/mol ¼ 1.0942 g
Content ¼ (1.0942 g/1.1025 g) 100% ¼ 99.25% (w/w)The requirement for content of
omeprazole in Ph.Eur. is 99.0–101.0% and therefore the raw material meets the
specification.
strong acid. The amount of the compound in the sample solution can then be determined as
the difference between the added strong acid and the excess of strong acid determined by
titration. An example of this is shown in Box 21.27.
For alkaline compounds (amines) that are hydrochlorides small amounts of extra
hydrochloric acid may be adsorbed to the compound. In a few cases a small amount of
the API may be present as the base. If such raw materials is titrated with base the result
obtained may be wrong, too high or too low, respectively. To avoid such errors, it is common
to add a minor amount of hydrochloric acid. The sample solution will then contain both the
protonated drug substance and the strong acid, and this mixture is then titrated with an
aqueous solution of a strong base. This is performed by potentiometric titration and a
392
Introduction to Pharmaceutical Chemical Analysis
Box 21.26 Quantitative determination of amitriptyline hydrochloride
by acid–base titration in aqueous environment according to Ph.Eur. 7th
edn (2011)
Assay: Dissolve 0.250 g in 30 ml of 96% ethanol R. Titrate with 0.1 M sodium hydroxide,
determining the endpoint potentiometrically (2.2.20).
1 ml of 0.1 M sodium hydroxide is equivalent to 31.39 mg of C20H24ClN.
Discussion of the procedure:
0.250 g raw material (BH þ ) is dissolved in 30 ml of ethanol. The solution is titrated with
0.1 M sodium hydroxide (OH). The endpoint is determined potentiometrically. In this
case the amitriptyline hydrochloride can be dissolved in water, but as the titration
progress the amitriptyline base is liberated and it is only very slightly soluble in water. To
avoid precipitation during titration it is necessary to dissolve the sample in ethanol. The
titrant is still an aqueous solution.
Chemical reaction:
BH þ þ OH ! B þ H2 O
The calculation is performed similarly to the calculation in Box 21.25.
Box 21.27 Quantitative determination of ephedrine by acid-base
titration in an acidic aqueous environment according to Ph.Eur. 7th
edn 2011
Assay: Dissolve 0.200 g in 5 ml of alcohol R and add 20.0 ml of 0.1 M hydrochloric acid.
Using 0.05 ml of methyl red solution R as indicator, titrate with 0.1 M sodium hydroxide
until a yellow color is obtained.
1 ml of 0.1 M hydrochloric acid is equivalent to 16.52 mg of C10H15NO.
Discussion of the procedure:
0.200 g raw material (B) is dissolved in 5 ml ethanol and then added to 20.0 ml of 0.1 M
HCl. 0.05 ml (one drop) of methyl red solution is added as an indicator and the solution is
titrated with 0.1 M sodium hydroxide (OH) until a yellow color is obtained.
Chemical reactions:
B þ 2HCl ! BH þ þ Cl þ HCl
HCl þ OH ! H2 O þ Cl
Chemical Analysis of Drug Substances
393
Example of calculation:
0.2099 g of raw material dissolved in 5 ml of ethanol and added to 20.00 ml of 0.1024 M
HCl is titrated with 7.75 ml of 0.0998 M NaOH. The molecular weight of ephedrine is
165.2 g/mol. The content of the raw material is calculated as follows:
The number of moles of consumed OH ¼ ð7:75ml=1000mlÞ 0:0998M
¼ 7:734 104 mol
The number of moles of HCl in excess ¼ 7:734 104 mol
Total number of moles of HCl added ¼ ð20:00ml=1000mlÞ 0:1024M
¼ 2:048 103 mol
The number of moles of HCl consumed ¼ moles of ephedrine ¼ 2:048 103 mol
7:734 104 mol ¼ 1:275 103 mol
Amount of ephedrine ðgÞ ¼ 1:275 103 mol 165:2g=mol ¼ 0:2106g
Content ¼ ð0:2106g=0:2099gÞ 100% ¼ 100:32% ðw=wÞ
The content of ephedrine, anhydrous should be 99.0–101.0% and the requirement is
therefore met.
titration curve is recorded. The curve will show two points of equivalence (inflexion). The
first point of inflexion corresponds to the strong acid and the second corresponds to the
protonated base. Thus the volume between the first and the second point of inflexion
corresponds to the amount of protonated base and is used for the calculation of content. An
example of this procedure is shown in Box 21.28.
21.6.2
Acid–Base Titration in a Non-Aqueous Environment
Non-aqueous acid-base titration is another type of titration often used for quantitative
determination of raw materials. As previously discussed in Chapter 5, this method is either
used: (i) when the drug is a very weak acid (pKa > 8) or a very weak base (pKb > 8), or (ii)
when the substance is poorly soluble in an aqueous environment or in polar organic solvents.
Non-aqueous titrations prescribed in Ph.Eur. are primarily used for quantitative determination of organic bases or of hydrochlorides, sulfates, phosphates and carboxylates of bases.
The raw material is dissolved in a non-aqueous solvent, and typically anhydrous acetic acid
and anhydrous formic acid are used. Sometimes acetic anhydride is also added. Acetic
anhydride is added to remove any remaining water in the sample solution, because water
may interfere with the determination. Acetic anhydride reacts with water according to the
following reaction:
ðCH3 COÞ2 O þ H2 O ! 2CH3 COOH
ð21:32Þ
394
Introduction to Pharmaceutical Chemical Analysis
Box 21.28 Quantitative determination of ephedrine hydrochloride by
acid–base titration in an acidic aqueous environment according to Ph.
Eur. 7th edn (2011)
Assay: Dissolve 0.150 g in 50 ml of 96% ethanol R and add 5.0 ml of 0.01 M hydrochloric
acid. Carry out a potentiometric titration (2.2.20) using 0.1 M sodium hydroxide. Read
the volume added between the two points of inflexion.
1 ml of 0.1 M sodium hydroxide is equivalent to 20.17 mg of C10H16ClNO.
Discussion of the procedure:
0.150 g of the raw material (BH þ Cl) is dissolved in 50 ml ethanol and is then added
5.0 ml of 0.01 M HCl. The sample solution is titrated with 0.1 M NaOH (OH) using
potentiometry. The volume of NaOH consumed between the two inflexion points
detected in the titration curve is determined and used for calculating the amount of
ephedrine hydrochloride. The content should be within 99.0–101.0% (w/w) based on
dried substance.
Chemical reactions:
HCl þ OH ! Cl þ H2 O
ðfirst point of inflexionÞ
BH þ þ OH ! B þ H2 O
ðsecond point of inflexionÞ
Example of calculation:
0.1487 g of raw material is dissolved and added 5.00 ml 0.0119 M HCl. The solution is
titrated using 0.60 ml of 0.0998 M NaOH to the first point of inflexion, and 7.95 ml to
the second point of inflexion. The molecular weight of ephedrine hydrochloride is
201.7 g/mol. The content of the raw material is calculated as follows:
OH (ml) between the two inflexion points ¼ 7.95 ml – 0.60 ml ¼ 7.35 ml
The number of moles OH consumed ¼ ð7:35 ml=1000 mlÞ 0:0998 M
¼ 7:734 104 mol
Moles ephedrine hydrochloride ¼ 7:734 104 mol
Amount of ephedrine hydrochloride ðgÞ ¼ 7:734 104 mol
201:7g=mol ¼ 0:1479g
Content of raw material ¼ ð0:1479g=0:1487gÞ 100% ¼ 99:49% ðw=wÞ
The raw material meets the requirement and the API is within specification if the loss on
drying is less than 0.5%.
Chemical Analysis of Drug Substances
395
Typically perchloric acid dissolved in acetic acid is used as the titrand. Perchloric acid is a
very strong acid and it reacts with acetic acid according to the following reaction:
CH3 COOH þ HClO4 ! CH3 COOH2þ þ ClO4
ð21:33Þ
The titrand reacts with basic compounds according to the following reaction:
B þ CH3 COOH2þ ! BH þ þ CH3 COOH
ð21:34Þ
The titrand CH3 COOH2þ is gradually added to the sample solution and the endpoint
determined potentiometrically or by using an indicator. An example of the use of nonaqueous acid–base titration is shown in Box 21.29.
Box 21.29 Quantitative determination of zopiclone by non-aqueous
acid–base titration according to Ph.Eur. 7th edn (2011)
Assay: Dissolve 0.300 g in a mixture of 10 ml of anhydrous acetic acid R and 40 ml of
acetic anhydride R. Titrate with 0.1 M perchloric acid, determining the endpoint
potentiometrically (2.2.20).
1 ml of 0.1 M perchloric acid is equivalent to 38.88 mg of C17H17ClN6O3.
Discussion of the procedure:
0.300 g of raw material (B) is dissolved in a mixture of 10 ml of anhydrous acetic acid and
40 ml of acetic anhydride. The sample solution is titrated with 0.1 M perchloric acid
dissolved in a mixture of glacial acetic acid and acetic anhydride (97 : 3). The endpoint is
determined potentiometrically. The content of zopiclone should be within 98.5 and
100.5% (w/w).
Chemical reaction:
B þ CH3 COOH2þ ! BH þ þ CH3 COOH
Example of calculation:
0.2996 g of the raw material is dissolved and then titrated with 7.45 ml of 0.1026 M
perchloric acid. The molecular weight of zopiclone is 388.8 g/mol. The content of the
raw material is calculated as follows.
Number of moles ¼ perchloric acid ð7:45 ml=1000 mlÞ 0:1026 M ¼ 7:644 104 mol
The number of moles of zopiclone ¼ 7:644 104 mol
Amount of zopiclone in g ¼ 7:644 104 mol 388:8g=mol ¼ 0:2972g
Content ¼ ð0:2972g=0:2996gÞ 100% ¼ 99:19% ðw=wÞ
The content of the raw material meets the requirement and the substance is within
specification.
396
Introduction to Pharmaceutical Chemical Analysis
Previously, hydrochlorides of weak bases like ephedrine hydrochloride (see Box 21.28)
were titrated in a non-aqueous medium using perchloric acid. But the protonated amine
could not be titrated directly with the strong acid. The sample was therefore treated with
mercury acetate before titration. The mercury acetate will react with the chloride ions
forming a soluble mercury chloride complex and a corresponding number of acetate ions
will be left in the solution:
Hg ðCH3 COOÞ2 þ 2Cl ! HgCl2 þ 2CH3 COO
ð21:35Þ
The acetate ions are now titrated with perchloric acid which in practice means that acetate
is titrated with the protonated acetic acid in the perchloric acid solution. The total reactions
are as follows:
2BH þ Cl þ Hg ðCH3 COOÞ2 ! 2BH þ þ HgCl2 þ 2CH3 COO
CH3 COO þ CH3 COOH2þ ! 2CH3 COOH
ð21:36Þ
ð21:37Þ
This type of non-aqueous titration has now been abandoned to avoid the used of mercury.
The titration of amine carboxylates are still performed according to the same reactions but
without the addition of mercury acetate. The titration of the anions corresponding to an
amine is of course an indirect determination of the active substance.
The titration of amine halogenide salts with perchloric acid using mercury acetate has
now been replaced either with an aqueous titration using sodium hydroxide or with a non
aqueous titration using formic acid as the solvent. Formic acid is a stronger acid than acetic
acid and is therefore in its protonated form able to titrate the very weak bases Cl and Br.
The reaction using perchloric acid dissolved in formic acid is therefore:
Cl þ HCOOH2þ ! HCl þ HCOOH
21.6.3
ð21:38Þ
Redox Titrations
Titrations based on redox reactions instead of acid–base reactions are also in use in Ph.Eur.
Such titrations are called redox titrations, and they are reviewed in Chapter 5. To a solution
of the raw material the titrant is added gradually until all the raw material has reacted (the
endpoint). If the raw material compound is in its reduced form an oxidizing agent is added
(the titrand) according to the following general equation:
Raw materialreduced form þ TitrantOxidized form ! Raw materialOxidized form þ TitrantReduced form
ð21:38Þ
The endpoint of the titration is detected as previously described either with a redox color
indicator or by amperometry. Box 21.30 shows an example of a quantitative determination
by means of a redox titration.
21.6.4
High Performance Liquid Chromatography
As previously discussed HPLC is used extensively in connection with test for related
substances. HPLC is now also being used increasingly for assay of drug substances.
Chemical Analysis of Drug Substances
397
Box 21.30 Quantitative determination of ferrous fumarate by redox
titration according to Ph.Eur. 7th edn (2011)
Assay: Dissolve with slight heating 0.150 g in 7.5 ml of dilute sulfuric acid R. Cool and
add 25 ml of water R. Add 0.1 ml of ferroin R. Titrate immediately with 0.1 M cerium
sulfate until the color changes from orange to light bluish-green.
1 ml of 0.1 M cerium sulfate is equivalent to 16.99 mg of C4H2FeO4.
Main procedure:
0.150 g of the raw material is dissolved in 7.5 ml of added sulfuric acid under slight
heating. The solution is cooled and 0.1 ml ferroin indicator and 25 ml of water are added.
The solution is titrated with 0.1 M cerium sulfate to the equivalence point where the color
changes from orange to pale blue-green.
Content (dried substance) should be within 93.0 and 101.0% (w/w).
The following reaction takes place:
Fe2 þ þ Ce4 þ ! Fe3 þ þ Ce3 þ
Example of calculation:
0.1511 g of the raw material is dissolved and titrated with 8.90 ml of 0.1002 M cerium
sulfate. The molecular weight of ferrous fumarate is 169.9 g/mol. The content of the raw
material is calculated as follows.
The number of moles of cerium sulfate ¼ ð8:90 ml=1000 mlÞ 0:1002 M
¼ 8:918 104 mol
The number of moles of ferrous fumarate ¼ 8:918 104 mol
Amount of ferrous fumarate in g ¼ 8:918 104 mol 169:9g=mol ¼ 0:1515g
Content of raw material ¼ ð0:1515g=0:1511gÞ 100% ¼ 100:27% ðw=wÞ
Especially UV spectrophotometry is being replaced for assays based on HPLC. HPLC is
much more specific than UV spectrophotometry and the accuracy and precision is also
considered to be better, but the HPLC equipment is more expensive. The principle for the
use of HPLC is that an accurate amount of raw materials is weighed and dissolved and then
diluted accurately in a volumetric flask. The sample solution is injected into an HPLC
system (see Chapter 13), and the area of the chromatographic peak is measured. The
quantitative determination is based on external calibration as discussed in Chapter 20 using
a reference solution with a known concentration. The peak area for the raw material to be
examined is compared to the peak area of a chemical reference substance (CRS) of the same
compound, and from this the content is determined. An example of a quantitative
determination using HPLC is shown in Box 21.31.
398
Introduction to Pharmaceutical Chemical Analysis
Box 21.31 Quantitative determination of simvastatin by HPLC
according to Ph.Eur. 7th edn (2011)
Assay: Liquid chromatography (2.2.29) as described in the test for related substances
with the following modification.
Injection: test solution and reference solution (c).
Calculate the percentage content of simvastatin from the declared content of simvastatin
CRS.
Related substances: Liquid chromatography (2.2.29). Prepare the solutions immediately
before use.
Solvent mixture: Mix 40 vol of 1.4 g/l solution of potassium dihydrogen phosphate R,
adjusted to pH 4.0 with phosphoric acid R, and 60 vol of acetonitrile R. Then, filter.
Test solution: Dissolve 75.0 mg of the substance to be examined in the solvent mixture
and dilute to 50.0 ml with the solvent mixture.
Reference solution (a): Dissolve 1.0 mg of simvastatin CRS and 1.0 mg of lovastatin
CRS (impurity E) in the solvent mixture and dilute to 50.0 ml with the solvent mixture.
Reference solution (b): Dilute 0.5 ml of test solution to 100.0 ml with the solvent mixture.
Reference solution (c): Dissolve 75.0 mg of simvastatin CRS in the solvent mixture and
dilute to 50.0 ml with the solvent mixture.
The chromatographic procedure:
. A stainless steel column 0.033 m long and 4.6 mm i.d. packed with end-capped
octadecylsilyl silica gel for chromatography R (3 mm);
. As mobile phase at a flow rate of 3.0 ml/min.
Gradient system:
Mobile Phase A: Acetonitrile and 0.1% (v/v) phosphoric acid (1:1).
Mobile Phase B: 0.1% (v/v) phosphoric acid in acetonitrile.
Gradient: See chart below.
% mobile phase B
% mobile phase A
0%
100 %
50 %
50 %
0%
100 %
4.5
8.0
Run time (min)
11.5 11.6 13.0
Chemical Analysis of Drug Substances
.
.
399
Detector: spectrophotometer at 238 nm;
Injection: 5 ml.
5 ml of reference solution (a) is injected. The test for related substances and assay are not
valid unless, in the chromatogram obtained, the resolution between the two peaks
corresponding to lovastatin and simvastatin is at least 4.0.
5 ml of the test solution and reference solution (c) are injected. Calculate the content of
simvastatin from the peak areas in the chromatograms obtained with test solution and
reference solution (c) and the declared content of simvastatin CRS.
Discussion of the procedure:
In order to perform the quantitative determination it is important to set up the HPLC
system. First the mobile phases are mixed according to the method description. The
HPLC pump is flushed with the mobile phases, and then the HPLC column is installed in
the HPLC system. When gradient elution is to be used it is recommended to analyze a
blank solution two or three times before the actual analysis is performed. It is important
to allow the HPLC system to equilibrate with the initial mobile phase for the system to be
in equilibrium before injection of the next sample. The first analysis to be performed is
the system suitability test (SST). Only if the data obtained (typically resolution, which
also is the case here) is within specifications the analytical work can be continued. If the
SST fails the HPLC system has to be adjusted before the analysis procedure is continued.
When the SST is passed the test solution and the reference solution can be injected. The
repeatability of the injections has to be in accordance with the requirements given in Ph.
Eur. Section 2.2.46 “Chromatographic Separation Techniques” and the content of the
raw material is calculated from the data obtained.
Example of SST and quantitative calculation:
Aweight of 0.0398 g of raw material is present in the test solution, an amount of 0.0401 g
of simvastatin CRS is in the reference solution (c). The purity of simvastatin CRS is
reported to 99.3%. The following results are obtained by HPLC the analysis.
System suitability test data.
SST
Simvastatin
Lovastatin
Retention time (min)
Peak width (tw; min)
2.63
1.71
0.19
0.15
Resolution (RS) between lovastatin and simvastatin is calculated as follows:
RS ¼
2ðtR2 tR1 Þ 2ð2:63 1:71Þ
¼ 5:4
¼
tw1 þ tw2
0:15 þ 0:19
400
Introduction to Pharmaceutical Chemical Analysis
Thus the SST is satisfied (should be > 4.0).
Quantitative analysis of simvastatin.
Peak area of Simvastatin (mean of n ¼ 5)
Reference solution (c)
Test solution
149 350
145 677
The contant of the raw material is calculated as follows:
Concentration of simvastatin in reference solution C
¼ ð0:0401 g=50:0 mlÞ ¼ 7:96 0993 104 g=ml ¼ 0:7964 mg=ml
Concentration of simvastatin in test solution
¼ ð145 677=149 350Þ 0:7964 mg=ml ¼ 0:7768 mg=ml
Amount of simvastatin in test solution
¼ 0:7768 mg=ml 50:0 ml ¼ 38:84 mg ¼ 0:0388 g
Content of raw material:
ð0:0388 g=0:0398 gÞ 100% ¼ 97:5%
The content shall be within 97.0–102.0% (dried substance) Thus the test is positive since
the resolution and retention time are consistent with the requirements of the HPLC
system and the purity is greater than 97%. Loss on drying is in the monograph set to
maximum 0.5%.
Comments on the procedure:
The sample solvent used for the reference and the sample solutions has approximately
the same composition as the mobile phase A. Looking at the retention time of simvastatin
and of lovastatin shows that both drugs are eluted during the isocratic part of the gradient.
These two substances are therefore not influenced by the gradient elution, but since the
HPLC system is also used for test for related substances, where more compounds can
vary considerably in chemical structure, it is prescribed as a gradient system. The last
part of the gradient from 8.0 to 11.5 min is used to elute any materials with high retention
before the next injection. The separation principle is reversed phase chromatography
using a C18 column, and detection is performed at 238 nm where the substances have
good light absorption.
Chemical Analysis of Drug Substances
21.6.5
401
UV Spectrophotometry
For some compounds it is not possible to perform quantitative determination by means of
titration because these compounds do not contain functional groups that can be easily
titrated. In such cases it is a possibility to make use of UV spectrophotometry for the
quantitative determination, but only if the substance absorp UV light. The principle of UV
spectrophotometry is described in Chapter 7. A known amount of raw material is dissolved
in a suitable solvent and the solution placed in a UV spectrophotometer. Since the
absorbance is directly proportional to the concentration of the dissolved sample (Beer’s
law) quantitative determination can be performed and the content of the substance to be
examined can be calculated. An example of the use of UV spectrophotometry for
quantitative determination is shown in Box 21.32.
Box 21.32 Quantitative determination of hydrocortisone by UV
spectrophotometry according to Ph.Eur. 7th edn (2011)
Assay: Dissolve 0.100 g in 96% ethanol R and dilute to 100.0 ml with the same solvent.
Dilute 2.0 ml of the solution to 100.0 ml with 96% ethanol R. Measure the absorbance
(2.2.25) at the maximum at 241.5 nm.
Calculate the content of C21H30O5, taking the specific absorbance to be 440.
Main procedure:
0.100 g of the raw material is dissolved in ethanol and diluted in a volumetric flask to
100.0 ml. 2.00 ml of this solution is diluted in a new volumetric flask to 100.0 ml with
ethanol. This solution is called the sample solution. Pure ethanol is filled into a quarts
cuvette which is placed in the UV spectrophotometer. The wavelength of UV light is set
at 241.5 nm and the absorbance is reset to 0.000 for ethanol. Then the sample solution is
filled into the cuvette and the absorbance of this solution is determined. The concentration of hydrocortisone is calculated based on the specific absorbance given as 440, and
from these data the content of the raw material is calculated.
Example of calculation:
0.0915 g of the raw material is dissolved in ethanol and diluted as prescribed. The
absorbance is measured to be 0.798. The content of the raw material is calculated as
follows, based on a specific absorbance (A) of 440 (1 cm, 1%) with a light path through
the cuvette of 1 cm:
The concentration of hydrocortisone in the test solution
¼ A=a b ¼ 0:798=½440 ðA; 1 cm; 1%Þ 1 cm ¼ 0:00181% ðw=vÞ
¼ 0:0181 mg=ml
The concentration of hydrocortisone in the first dilution of the sample
¼ ð100:0 ml=2:00 mlÞ 0:0181 mg=ml ¼ 0:9050 mg=ml
402
Introduction to Pharmaceutical Chemical Analysis
Amount of hydrocortisone in the first dilution of the sample
¼ 0:9050 mg=ml 100:0 ml ¼ 90:50 mg ¼ 0:0905 g
The content of the raw material is therefore:
ð0:0905 g=0:0915 gÞ 100% ¼ 98:9%
The content of hydrocortisone has to be between 97.0 and 103.0%, and the content is thus
within the requirement.
Comments on the procedure:
Ethanol is used as the sample solvent because hydrocortisone dissolves easily in this
solvent, and because it has no absorbance at 241.5 nm. The weighed amount of sample is
100 mg to ensure that uncertainty is kept to a minimum. The raw material is diluted to
obtain an absorbance within the range of 0.2–0.8 where the highest precision is obtained.
21.7 How to Control Compounds for Which no Pharmacopoeia
Monograph Exists
Until now focus has been on how to conduct control of pharmaceutical raw materials for
which monographs (standards) exist in a pharmacopoeia. As previously mentioned, you will
find monographs for most of the excipients in Ph.Eur. and USP, and a large number of
monographs on active substances (APIs) are also collected here. Newer drug substances are
often protected by patents, and they are not included in the pharmacopoeias. The reason for
this is that the substances are only used by one manufacturer, namely, the company that
holds the patent on the compound. It is of little relevance to publish monographs for such
drug substances until the patent expires. The manufacturer using such a substance will for
internal purposes have to develope a monograph on the substance for raw material control,
and this shall be submitted to the authorities as part of the documentation for marketing
authorization. These monographs are designed in the same way as those found in the Ph.Eur.
When the patent expires, it is common that the monograph is included in the Ph.Eur, because
other manufacturers will appear on the market with the same substance. In connection to Ph.
Eur. guidelines (technical guide) for how new monographs should be prepared is published
in order to secure that the monographs are designed in a standardized way.
21.8 How are Ph.Eur. and USP Updated?
Ph.Eur. 7th edn (2011) is the latest edition of the European Pharmacopoeia. Similarly USP
34 – NF 29 is the latest edition of the United States Pharmacopoeia. As the suffix implies it
also include standards given in the National Formulary 29.
Chemical Analysis of Drug Substances
403
The pharmacopoeias are updated with regular intervals and in between the new full
editions a number of supplements are published in order to keep everything fully updated.
The supplements contain new monographs on new substances as well as correction to
existing monographs and chapters. Thus it is important to consult the latest supplement of a
pharmacopoeia to ensure that the latest updates are taken into consideration.
The number of monographs on compounds has increased significantly in recent years,
and the work has thus become considerably more extensive. In addition to this, the way of
testing has changed somewhat over the years primarily due to developments in analytical
chemical technology. In the older editions a larger number of very simple tests were used,
and the reason for this was that much production was done on a small scale in pharmacies
and in small businesses. In recent years, the majority of drug production takes place in large
industrial companies where the supply of more sophisticated and modern analytical
chemical equipment is a necessity. Accordingly, the pharmacopoeias have gradually
adopted more advanced methods, and especially HPLC has become a very important
tool for assay and purity testing. It is now normal to use HPLC for the testing of related
substances, where TLC previously was used. HPLC is now also increasingly being used for
quantitative determinations. One of the advantages is that testing for related substances and
doing quantitative determination can be performed in one operation. The disadvantage
compared to the titrations is the need for a reference substance of each drug substance to be
available, while the titration only requires a suitable primary standard. For identification
purposes IR has become the instrumental technique of choice and it is nearly always the
primary test for identification. In addition TLC and some color reactions are still used,
especially for drug substances handled in pharmacies where more simple tests are needed.
The pharmacopoeias will also in the future be updated in order to keep up with general
developments.
22
Chemical Analysis of Final
Pharmaceutical Products
Quality control of the final pharmaceutical product is an important part of the entire
pharmaceutical production. Quality control of final pharmaceutical products includes both
microbiological tests, pharmaceutical tests, and tests based on chemical analysis (chemical
tests). This chapter directs our focus towards the latter subject, namely the chemical tests of
final pharmaceutical products. The general rules and requirements are discussed, and the
different chemical tests that are normally applied to the final pharmaceutical products are
discussed in detail. Because a large number of different methods are used in the pharmaceutical industry, this chapter cannot cover them all. However, with some selected examples,
which are discussed in detail, this chapter is intended to give a basic understanding.
22.1 Quality Control of Final Pharmaceutical Products
Chapter 21 has discussed in detail how the active pharmaceutical ingredients (APIs) and
pharmaceutical excipients are quality controlled according to the European Pharmacopoeia
and related pharmacopoeias. This is important to ensure a high quality of the ingredients,
which in turn is of the highest importance for the safety of the patients. In addition to quality
control of the ingredients, quality control of the final pharmaceutical product is also of the
highest importance for the safety of patients.
Quality control of the final pharmaceutical product can in principle be divided into three
different areas, according to the nature of the experimental methods in use:
.
Microbiological testing;
Pharmaceutical testing;
. Chemical testing.
.
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
406
Introduction to Pharmaceutical Chemical Analysis
In general one can say that the microbiological tests are used to check the final pharmaceutical product with respect to different bacteria and related microorganisms. Microbiological
tests are very important both for sterile and nonsterile products, but because they rely on the
concepts of microbiology these tests are outside the scope of this book. However, it is
important to emphasize that microbiological tests are a very important subject in pharmaceutical quality control, a very important subject of the pharmaceutical curriculum, and
their teaching is normally given in courses related to microbiology.
Pharmaceutical testing is intended to check the quality of the dosage form itself. The
pharmaceutical testing is based on physical methods, and these are normally discussed in
courses in pharmaceutics. Pharmaceutical tests are also outside the scope of this book, but
again it is important to emphasize that these tests are also a highly important part of the
quality control.
Chemical tests are intended to check the quality of the final pharmaceutical product in
terms of the chemical composition. Such chemical tests rely to a large extent on spectroscopic and chromatographic techniques as well as titration techniques. These techniques
have been discussed earlier in this book. This chapter directs our focus towards how to use
these techniques for the chemical testing of final pharmaceutical products.
22.2 Monographs and Chemical Testing
In Chapter 21, you learned that chemical testing of different active pharmaceutical
ingredients (APIs) and pharmaceutical excipients has to be done in compliance with the
individual monographs in the European Pharmacopoeia when intended for the European
market, or in compliance with the United States Pharmacopeia when intended for the United
States market. Similar requirements hold for chemical testing of the final pharmaceutical
products, although the European Pharmacopoeia is not very specific on this subject, as
discussed initially in this section.
The European Pharmacopoeia has no individual monographs for the different final
pharmaceutical products, but contains general monographs for the dosage forms. The
chemical testing of products for the European market has to be in compliance with these
general monographs. Examples of dosage forms are capsules, parenteral preparations, and
tablets. The European Pharmacopoeia monograph for capsules is shown in Box 22.1.
The general monographs in the European Pharmacopoeia for dosage forms are composed
of the following sections:
.
Definition;
Production;
. Tests.
.
The first section is the Definition. This section first gives a clear definition of the dosage
form, and a general description about the production. This is exemplified in Box 22.1 for
capsules. Also, the Definition section gives an overview of the excipients typically used in
the dosage form. This is also illustrated in Box 22.1 for capsules. Furthermore, the
Definition section gives the requirements for the material of the containers intended for
the final pharmaceutical product. All pharmaceutical products have to comply with the
requirements stated in the section Materials used for the manufacture of containers.
Chemical Analysis of Final Pharmaceutical Products
407
Box 22.1 European Pharmacopoeia monograph for capsules
CAPSULES
Capsulae
The requirements of this monograph do not necessarily apply to preparations that are
presented as capsules intended for use other than by oral administration. Requirements
for such preparations may be found, where appropriate, in other general monographs,
for example Rectal preparations (1145) and Vaginal preparations (1164).
DEFINITION
Capsules are solid preparations with hard or soft shells of various shapes and capacities,
usually containing a single dose of active substance(s). They are intended for oral
administration.
The capsule shells are made of gelatin or other substances, the consistency of which
may be adjusted by the addition of substances such as glycerol or sorbitol. Excipients
such as surface-active agents, opaque fillers, antimicrobial preservatives, sweeteners,
coloring matter authorized by the competent authority and flavouring substances may be
added. The capsules may bear surface markings.
The contents of capsules may be solid, liquid or of a more paste-like consistency. They
consists of one or more active substances with or without excipients such as solvents,
diluents, lubricants and disintegrating agents. The contents do not cause deterioration
of the shell. The shell, however, is attacked by the digestive fluids and the contents
are released.
Where applicable, containers for capsules comply with the requirements of Materials
used for the manufacture of containers (3.1 and subsections) and Containers (3.2 and
subsections).
Several categories of capsules may be distinguished:
– Hard capsules;
– Soft capsules;
– Gastro-resistant capsules;
– Modified-release capsules;
– Cachets.
PRODUCTION
In the manufacture, packaging, storage and distribution of capsules, suitable measures
are taken to ensure their microbial quality; recommendations on this aspect are provided
in the text on Microbiological quality of pharmaceutical preparations (5.1.4).
TESTS
Uniformity of dosage units. Capsules comply with the test for uniformity of dosage
units (2.9.40) or, where justified and authorized, with the tests for uniformity of content
408
Introduction to Pharmaceutical Chemical Analysis
and/or uniformity of mass shown below. Herbal drugs and herbal drug preparations
present in the dosage form are not subject to the provisions of this paragraph.
Uniformity of content (2.9.6). Unless otherwise prescribed or justified and authorized,
capsules with a content of active substance less than 2 mg or less than 2 per cent of the fill
mass comply with test B for uniformity of content of single-dose preparations. If the
preparation has more than one active substance, the requirement applies only to those
ingredients which correspond to the above conditions.
Uniformity of mass (2.9.5). Capsules comply with the test for uniformity of mass of
single-dose preparations. If the test for uniformity of content is prescribed for all the
active substances, the test for uniformity of mass is not required.
Dissolution. A suitable test may be carried out to demonstrate the appropriate release of
the active substance(s), for example one of the tests described in Dissolution test for solid
dosage forms (2.9.3).
Where a dissolution test is prescribed, a disintegration test may not be required.
STORAGE
Store at a temperature not exceeding 30 C.
For capsules, this is illustrated in Box 22.1. The reason for these requirements is to make
sure that the final pharmaceutical product is not deteriorated in contact with the container.
The Materials used for the manufacture of containers section contains monographs for the
materials used for containers, like polyvinyl chloride (PVC), polyethylene (PE), and
polypropylene (PP). These monographs are structured much in the same way as other
monographs in the pharmacopoeia, and include sections about definition, production,
identification, and tests. Although very important, this aspect is not discussed further in
this textbook.
The Definition section of the general monographs for dosage forms also gives the
requirements for the container itself. All dosage forms have to comply with requirements
stated in the section Containers. The reason for this is to make sure that the final
pharmaceutical product is not deteriorated by sun light, moisture, air, contamination,
leakage, or mechanical stress, or by adsorption of ingredients to the container wall. For
capsules, this is illustrated in Box 22.1.
The next section is Production. This section gives precautions related to the production of
the specific dosage form. In addition, the Production section gives the requirements for
different tests to make sure that the production process has been successful. The tests can be
pharmaceutical tests to ensure the technical quality of the product, or microbiological tests
to ensure the microbiological quality of the product. In common for all dosage forms,
production, packing, storage, and distribution have to comply with the requirements stated
in the section Microbiological quality of pharmaceutical preparations. For capsules, this is
exemplified in Box 22.1. For sterile preparations, the production also has to comply with the
requirements stated in the section Methods of preparation of sterile products. For some
products, antimicrobial preservatives are added, and in those cases, the Production section
Chemical Analysis of Final Pharmaceutical Products
409
explains that the product has to be tested according to the section Efficacy of microbial
preservation, to make sure that the efficiency of the antimicrobial preservatives is sufficient.
As stated earlier, microbiological tests are not discussed in detail in this textbook.
The final section is Tests. The purpose of the tests is mainly to ensure the pharmaceutical
quality of the product. The prescribed tests are quite different from dosage form to dosage
form. For tablets and capsules as examples, the tests are typically intended to check that the
dosage form is disintegrated and the active ingredient is released in an appropriate way.
Also, for tablets and capsules, tests are performed to check that the active pharmaceutical
ingredient is equally distributed in each dosage unit. For parenteral preparations, as another
example, tests are performed to check that the product does not contain contamination by
particulate matter, and that the product is sterile.
In contrast to the European Pharmacopoeia, both the British Pharmacopoeia and the
United States Pharmacopeia include individual monographs for final pharmaceutical
products. Thus, chemical testing of pharmaceutical products intended for the United
Kingdom and United States markets has to be in compliance with those individual
monographs. One example of such an individual monograph is shown in Box 22.2 for
paracetamol tablets from the British Pharmacopoeia.
Box 22.2 British Pharmacopoeia monograph for paracetamol tablets
Paracetamol Tablets
Action and use
Analgesic; antipyretic.
DEFINITION
Paracetamol Tablets contain Paracetamol.
The tablets comply with the requirements stated under Tablets and with the following
requirements.
Content of paracetamol, C8H9NO2
95.0 to 105.0% of the stated amount.
IDENTIFICATION
Extract a quantity of the powdered tablets containing 0.5 g of Paracetamol with 20 mL of
acetone, filter, evaporate the filtrate to dryness and dry at 105 C. The residue complies
with the following tests.
A. The infrared absorption spectrum, Appendix II A, is concordant with the reference
spectrum of paracetamol (RS 258)
B. Boil 0.1 g with 1 mL hydrochloric acid for 3 minutes, add 10 mL of water and cool;
no precipitate is produced. Add 0.05 mL of 0.0167 M potassium dichromate; a violet
color is produced slowly which does not turn red.
C. Melting point, about 169 C, Appendix V A.
410
Introduction to Pharmaceutical Chemical Analysis
TESTS
Dissolution
Comply with the requirements for the Monographs of the British Pharmacopoeia in the
dissolution test for tablets and capsules, Appendix XII B1, using Apparatus 2. Use as the
medium 900 mL of phosphate buffer pH 5.8 and rotate the paddle at 50 revolutions per
minute. Withdraw a sample of 20 mL of the medium and filter. Dilute the filtrate with
0.1 M sodium hydroxide to give a solution expected to contain about 0.00075% w/v of
Paracetamol. Measure the absorbance of this solution, Appendix II B, at the maximum at
257 nm using 0.1 M sodium hydroxide in the reference cell. Calculate the total content of
paracetamol, C8H9NO2, in the medium taking 715 as the value of A (1%, 1 cm) at the
maximum at 257 nm.
Related substances
Carry out the method for liquid chromatography, Appendix III D, using the
following solutions. Prepare the solutions immediately before use and protect
from light. For solution (1) disperse a quantity of powdered tablets containing
0.2 g of Paracetamol in 8 mL of the mobile phase with the aid of ultrasound, add
sufficient mobile phase to produce 10 mL, mix well and filter. For solution (2) dilute
1 volume of solution (1) to 20 volumes with mobile phase and dilute 1 volume of this
solution to 20 volumes with mobile phase. Solution (3) contains 0.002% w/v each of
4-aminophenol and paracetamol BPCRS in the mobile phase. Solution (4) contains
0.00002% w/v of 40 -chloroacetanilide in the mobile phase.
The chromatographic procedure may be carried out using (a) a stainless steel column
(25 cm 4.6 mm) packed with octylsilyl silica gel for chromatography (5 mm) (Zorbax
Rx C8 is suitable), (b) as the mobile phase with a flow rate of 1.5 mL per minute, at a
temperature of 35 C, a mixture of 250 volumes of methanol containing 1.15 g of a 40%
v/v solution of tetrabutylammonium hydroxide with 375 volumes of 0.05 M disodium
hydrogen orthophosphate and 375 volumes of 0.05 M sodium dihydrogen orthophosphate and (c) a detection wavelength of 245 nm.
The test is not valid unless, in the chromatogram obtained for solution (3), the
resolution factor between the two principal peaks is at least 4.0.
Inject solution (1) and allow the chromatography to proceed for 12 times the retention
time of the principal peak. In the chromatogram obtained with solution (1) the area of any
peak corresponding to 4-aminophenol is not greater than the area of the corresponding
peak in solution (3) (0.1%), the area of any peak corresponding to 40 -chloroacetanilide is
not greater than the area of the principal peak in solution (4) (10 ppm) and no other
impurity is greater than the area of the principal peak obtained with solution (2) (0.25%).
ASSAY
Weigh and powder 20 tablets. Add a quantity of the powder containing 0.15 g of
Paracetamol to 50 mL of 0.1 M sodium hydroxide, dilute with 100 mL of water, shake for
15 minutes and add sufficient water to produce 200 mL. Mix, filter and dilute 10 mL of
the filtrate to 100 mL with water. Add 10 mL of the resulting solution to 10 mL of 0.1 M
Chemical Analysis of Final Pharmaceutical Products
411
sodium hydroxide, dilute to 100 mL with water and measure the absorbance of the
resulting solution at the maximum at 257 nm, Appendix II B. Calculate the content of
C8H9NO2 taking 715 as the value of A(1%, 1 cm) at the maximum at 257 nm.
STORAGE
Paracetamol Tablets should be protected from light.
The first part of the monograph gives a definition of the final pharmaceutical product. In
the case of paracetamol tablets, it is first stated that paracetamol tablets contain paracetamol
as the active pharmaceutical ingredient. Also, the monograph states that the tablets should
comply with the requirements defined in the general monograph for tablets. Finally, the
definition section defines the tolerance for the content of active pharmaceutical ingredient.
In the case of paracetamol tablets, the content should be within the range 95.0–105.0% of the
stated amount.
The next part of the monograph describes the Identification procedures. These are
intended to identify the active pharmaceutical ingredient in the final pharmaceutical
product, or to say it in other words, to make sure that the final pharmaceutical product
contains the correct active pharmaceutical ingredient. In the case of paracetamol tablets, the
tablets are first powdered, the powder is extracted with acetone, and the acetone is
evaporated to dryness. The residue, which is the active pharmaceutical ingredient, is either
tested by IR spectrophotometry, by a simple color test, or by melting point for positive
identification. Different examples of identification procedures for active pharmaceutical
ingredients are discussed in detail later in this chapter.
The next section is Tests, which describes different test methods prescribed for the final
pharmaceutical product. Often, this section contains a test for related substances, where the
final pharmaceutical product is tested for traces of degradation products from the active
pharmaceutical ingredient, or other trace level substances related to the active pharmaceutical ingredient. This is also the case for paracetamol tablets in Box 22.2, where the test is
based on HPLC of tablet powder dispersed in mobile phase and subsequently filtered. The
Test section can also contain other types of test procedures. For solid formulations, like
tablets and capsules, the Tests section may, among others, include tests for Dissolution and
Uniformity of content. The test for dissolution is intended to check that the active
pharmaceutical ingredient is released with an appropriate rate when dissolved, or to say
it in other words, to make sure that the active pharmaceutical ingredient is released in the
gastrointestinal system. Uniformity of content is intended to check that the content of active
pharmaceutical ingredient does not vary too much from tablet to tablet. Both dissolution and
uniformity of content are based on chemical methods and are discussed briefly later in this
chapter. For liquid formulations, like injections and oral solutions, the Tests section may,
among others, include tests for clarity, pH, light absorption, and refractive index. These tests
are intended to check that the final pharmaceutical product has the correct pH value and that
there is no nondissolved matter in the product.
The final section is Assay. This section describes the procedure for quantitative analysis of
the active pharmaceutical ingredient in the final pharmaceutical product. For paracetamol
tablets (Box 22.2), the assay is based on UV spectroscopy.
412
Introduction to Pharmaceutical Chemical Analysis
As discussed above, final pharmaceutical products intended for the United Kingdom or
United States markets have to comply with the corresponding monographs published in the
British Pharmacopoeia or in the United States Pharmacopeia. Thus, the chemical testing
follows the monographs very closely. For the rest of Europe, the pharmaceutical industry
does chemical testing of final pharmaceutical products according to their own specifications, but these specifications are very often similar to the monographs published in the
British Pharmacopoeia and the United States Pharmacopeia. Therefore, the monographs
from the British Pharmacopoeia will be used as examples during the rest of this chapter to
illustrate and discuss in detail the chemical testing of final pharmaceutical products. It is
important to emphasize that this discussion is not complete, as the pharmaceutical industry
may use additional tests in their quality control.
22.3 Identification of the Active Pharmaceutical Ingredient
Identification tests are intended to identify the active pharmaceutical ingredient in the final
pharmaceutical product. In some cases, particular excipients may also be included in the
identification tests. This may for instance be the case for antimicrobial preservatives. The
major analytical techniques used for identification testing are IR spectrophotometry, UV
spectrophotometry, high-performance liquid chromatography (HPLC), and thin-layer
chromatography (TLC). The use of these techniques are illustrated below with selected
examples. In addition, color reactions may be used to some extent, but these are not
discussed in this textbook as they are less in use in modern pharmaceutical laboratories.
IR spectrophotometry is frequently in use, and the fundamentals of IR have been
discussed in Chapter 8. IR is a very specific method for the identification of active
pharmaceutical ingredients in pharmaceutical products. An IR spectrum of the active
pharmaceutical ingredient that gives exact match with a similar spectrum from chemical
reference substance of the same drug, gives a highly reliable identification. This is the main
advantage of using IR for identification. However, since excipients may interfere in the IR
spectrum, the active pharmaceutical ingredient has to be isolated from the excipients before
IR identification. Thus, sample preparation and extraction is required. The procedures used
for this depends on the type of formulation and on the complexity of the formulation. In
common however, the procedures end up with a dry residue of the active pharmaceutical
ingredient, from which an IR spectrum can be obtained. A few typical examples are
illustrated and discussed in Boxes 22.3, 22.4, and 22.5. Read carefully through these
examples, and from the discussion of the principles, you should understand each step in
the procedure.
UV spectrophotometry is also frequently used for the identification of active pharmaceutical ingredients. The fundamentals of UV have been discussed in Chapter 7. UV is less
specific than IR, but is still acceptable for the identification of active pharmaceutical
ingredients. Typically, a UV spectrum from the active pharmaceutical ingredient is
recorded, and one or two absorption maxima should be at specified wavelengths for
positive identification. In addition, the absorbance at absorption maximum can be measured
and should be within a specified range for positive identification. As for IR, excipients
may interfere during the UV measurement, and in such cases, the active pharmaceutical
Chemical Analysis of Final Pharmaceutical Products
413
Box 22.3 Identification of paracetamol in paracetamol tablets by IR
spectrophotometry (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
OH
O
N
H
Molecular weight: 151.2 g/mol
Acid/base:
Acidic (weak)
pKa:
9.5
Log P:
0.87
Procedure:
Extract a quantity of the powdered tablets containing 0.5 g of paracetamol with 20 ml of
acetone, filter, evaporate the filtrate to dryness and dry at 105 C. The infrared absorption
spectrum of the dried residue is concordant with the reference spectrum of paracetamol.
Discussion of the procedure:
During extraction with acetone, paracetamol is isolated from the tablet excipients, as the
latter are almost insoluble in acetone, whereas paracetamol is soluble in acetone. The
tablet excipients are then removed by filtration. Since IR spectrophotometry preferably
is carried out in solid state, the extraction solvent is evaporated to dryness to give a
residue of pure paracetamol. This residue is dried at 105 C to remove traces of water,
which can interfere seriously during the IR spectrophotometry, and an IR spectrum can
be obtained on the residue. The IR spectrum from this is then compared with a reference
spectrum of paracetamol, and if they are identical, paracetamol is positively identified in
the tablets.
Transmittance (%)
100
80
60
40
20
2000
1800
1600
1400
1200
1000
Wavenumber (cm-1)
800
600
400
414
Introduction to Pharmaceutical Chemical Analysis
How much tablet powder should be used?
Tablets containing 500 mg paracetamol is to be analyzed. Ten tablets are collected and the
total weight of the 10 tablets is 5.6322 g. The average weight of each tablet (containing
500 mg of paracetamol as specified content) is: 5.6322 g/10 tablets ¼ 0.5632 g.
Tablet powder corresponding to 500 mg of paracetamol is equivalent to one tablet,
which corresponds to 0.5632 g tablet powder. Thus, 0.5632 g tablet powder should be
used for the identification. However, since this is not a quantitative method, exact
weighing of 0.5632 g tablet powder is not critical.
Box 22.4 Identification of fluoxetine in fluoxetine hydrochloride oral
solution by IR spectrophotometry (adopted from British Pharmacopoeia
2009)
Chemical properties of the active pharmaceutical ingredient:
F
F
F
N
H
Molecular weight:
Acid/base:
pKa:
Log P:
O
309.3 g/mol
Basic
9.8
4.2
Procedure:
To a volume of the oral solution containing the equivalent of 0.1 g fluoxetine add 5 ml of
4 M sodium hydroxide, extract with 10 ml of dichloromethane, filter the dichloromethane layer and evaporate the filtrate to dryness. The infrared absorption spectrum of
the residue is concordant with the reference spectrum of fluoxetine.
Discussion of the procedure:
The oral solution, which has a pH of about 6, is first made alkaline to suppress the
ionization of the basic active pharmaceutical ingredient, and then this neutral substance
is extracted into dichloromethane by liquid–liquid extraction. The dichloromethane
extract is filtered to remove any particulate matter, and the filtrate is evaporated to
dryness to remove the solvent. The residue is the active pharmaceutical ingredient, from
which an IR spectrum can be obtained. The IR spectrum from this is compared with a
Chemical Analysis of Final Pharmaceutical Products
415
reference spectrum of fluoxetine, and if they are identical, fluoxetine is positively
identified in the oral solution.
Transmittance (%)
100
80
60
40
20
2000
1800
1600
1400
1200
1000
800
600
400
Wavenumber (cm-1)
How much oral solution should be used?
Oral solution containing 5 mg/ml fluoxetine is to be analyzed. 0.1 g (100 mg) of
fluoxetine is equivalent to:
100 mg at 5:0 mg=ml ¼ 20 ml oral solution
Thus, 20 ml oral solution should be used for the identification. However, since this is
not a quantitative method, collection of the oral solution by an exact transfer pipette is
not critical.
Box 22.5 Identification of mupirocin in mupirocin calcium nasal
ointment by IR-spectrophotometry (adopted from British
Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
OH
OH
O
OH
O
O
O
O
OH
Molecular weight: 500.6 g/mol
Acid/base:
Acidic
pKa:
4.8
Log P:
2.4
Procedure:
Disperse a quantity of the nasal ointment containing the equivalent of 100 mg of
mupirocin calcium in a mixture of 20 ml of dichloromethane and 4 ml of 0.2 M sodium
416
Introduction to Pharmaceutical Chemical Analysis
borate buffer pH 8.4, with shaking, and centrifuge at 3000 rpm until separate layers are
obtained. Remove and retain the upper aqueous layer and re-extract the lower layer with
a further 4 ml of 0.2 M sodium borate buffer pH 8.4. Combine the aqueous extracts, add
10 ml of dichloromethane to the combined extracts and centrifuge at 3000 rpm. Remove
the upper aqueous layer and adjust the pH to about 4 with 1 M hydrochloric acid (a slight
cloudiness may be observed), add 10 ml of dichloromethane and centrifuge. Separate
and retain the lower layer; add a further 10 ml dichloromethane to the upper layer and
centrifuge. Combine the dichloromethane extracts and evaporate to dryness at room
temperature under a current of nitrogen; a jelly-like residue is obtained. The infrared
absorption spectrum of the dried residue is concordant with the reference spectrum
of mupirocin.
Discussion of the procedure:
Ointment is a complex formulation where the active pharmaceutical ingredient is
distributed in a lipid vehicle. Thus, the formulation is not soluble in water. To disperse
the vehicle, and to get access to mupirocin, dichloromethane is added to the sample.
Borate buffer is then added, resulting in a two-phase system. The fatty ointment vehicle
remains in the dichloromethane phase, whereas the acidic drug (mupirocin) is extracted
into the slightly basic borate buffer. In this way, mupirocin is isolated from the bulk
formulation. Extraction with borate buffer is accomplished two times to ensure an
efficient extraction of mupirocin. The two portions of borate buffer (containing
mupirocin) are combined, and extracted with a new portion of dichloromethane.
This is accomplished to further purify the aqueous phase, to make sure that no excipients
are present in the aqueous extract. Finally, the aqueous phase is acidified, which suppress
the ionization of mupirocin, and mupirocin is extracted into dichloromethane. This
extraction is repeated with a second portion of dichloromethane. The two dichloromethane extracts are combined, and the solvent is evaporated. The residue is the active
pharmaceutical ingredient, from which an IR spectrum can be obtained. The IR spectrum
from this is compared with a reference spectrum of mupirocin, and if they are identical,
mupirocin is positively identified in the nasal ointment.
Transmittance (%)
100
80
60
40
20
2000
1800
1600
1400
1200
1000
Wavenumber (cm-1)
800
600
400
Chemical Analysis of Final Pharmaceutical Products
417
How much ointment should be used?
Ointment containing 2% (w/w) mupirocin calcium is to be analyzed. 100 mg (0.100 g)
mupirocin calcium is equivalent to the following amount of ointment:
½ð100%Þ=ð2%Þ 0:100 g 5:0 g
Thus, 5.0 g ointment should be used for the identification. However, since this is not a
quantitative method, exact weighing of 5.0000 g ointment is not critical.
ingredient has to be isolated from the excipients. Thus, some sample preparation is required.
The procedures differ from product to product, but generally the procedures are completed
by dissolution of the isolated active pharmaceutical ingredient in an aqueous sample or in
some polar organic solvent like methanol or ethanol. Aqueous solutions or solutions in
methanol or ethanol are well suited for UV spectrophotometry because of low background
absorbance from the solvent. Some examples on the use of UV for identification are
discussed in Boxes 22.6, 22.7, and 22.8.
HPLC is also a very popular method for identification. The fundamentals of HPLC have
been discussed in Chapter 13. Liquid products can be injected more or less directly in the
HPLC system for analysis, whereas solid or semi-solid products have to be dissolved in
mobile phase before HPLC. The retention time of the active pharmaceutical ingredient is
measured, and this retention time is compared with the retention time for a chemical
Box 22.6 Identification of diazepam in diazepam tablets by UV
spectrophotometry (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
H
N
Cl
Molecular weight: 284.7 g/mol
Acid/base:
Basic (weak)
pKa:
2.9
Log P:
3.1
N
418
Introduction to Pharmaceutical Chemical Analysis
Procedure:
Weigh and powder 20 tablets. To a quantity of the powder containing 10 mg of diazepam
add 5 ml of water, mix and allow to stand for 15 min. Add 70 ml of a 0.5% w/v solution of
sulfuric acid in methanol, shake for 15 min, add sufficient of the methanolic sulfuric acid
to produce 100 ml and filter. Dilute 10 ml of the filtrate to 50 ml with the same solvent and
measure the absorbance of the resulting solution in the range 230–350 nm. The solution
shows two maxima, at 242 and 284 nm.
Discussion of the procedure:
First, the tablets are powdered to access the total amount of active pharmaceutical
ingredient (diazepam). 20 tablets are selected to represent the sample, but only a small
amount of the resulting powder, corresponding to 10 mg of diazepam, is required for the
identification. First the tablet powder is suspended in pure water, and thereafter acidified
methanol is added to dissolve diazepam. Diazepam is soluble in methanol, but the
solubility is improved by acidifying methanol with sulfuric acid. Most excipients from
the tablets are not dissolved in acidic methanol, and they are removed by the subsequent
filtration. The methanolic solution containing the active pharmaceutical ingredient is
further diluted in order not to overload the UV instrument. This final solution is filled into
a quarts cuvette, and the UV spectrum is recorded between 230 and 350 nm. Absorbance
maxima should be observed at 242 and 284 nm for positive identification of diazepam.
How much tablet powder should be used?
Tablets containing 5 mg diazepam is to be analyzed. 20 tablets are collected and the total
weight is 2.6656 g. The average weight of each tablet (containing 5 mg of diazepam as
specified content) is:
2:6656 g=20 tablets ¼ 0:1330 g
UV absorbance
242
284
200
250
300
350
Wavelength (nm)
400
Chemical Analysis of Final Pharmaceutical Products
419
Tablet powder corresponding to 10 mg of diazepam is equivalent to two tablets, which
corresponds to:
0:1330 g 2 tablets 0:27 g tablet powder
Thus, 0.27 g tablet powder should be used for the identification. However, since this is
not a quantitative method, exact weighing of 0.2700 g tablet powder is not critical.
Box 22.7 Identification of flupentixol decanoate in flupentixol decanoate
injection by UV spectrophotometry (adopted from British
Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
O
N
N
F
F
F
S
Molecular weight: 588.8 g/mol
Acid/base:
Basic
Procedure:
Dilute a volume of the injection to contain 0.2% (w/v) of flupentixol decanoate. Dilute
1 vol of this solution to 2 vol with ethanol (96%). The light absorption in the range
205–300 nm of the resulting solution exhibits two maxima at 230 and 264 nm.
Discussion of the procedure:
Injections are normally formulations of low complexity, and can often be analyzed by
UV spectrophotometry more or less directly. In this case, the injection is diluted to ensure
that the UV instrument is not overloaded, or to state it differently, to ensure that
absorbances are in the range 0.2 to 0.8 (as discussed in Chapter 7). This final solution is
420
Introduction to Pharmaceutical Chemical Analysis
filled into a cuvette, and the UV spectrum is recorded between 205 and 300 nm.
Absorbance maxima should be observed at 230 and 264 nm for positive identification
of flupentixol.
UV absorbance
230
264
200
250
300
350
400
Wavelength (nm)
How much injection should be used?
Injection containing 20 mg/ml flupentixol decanoate is to be analyzed. Injection
containing 0.2% (w/v) (or 2 mg/ml) of flupentixol decanoate is equivalent to the original
injection diluted by a factor of 10. Thus, the injection should be diluted 10 times with
water prior to analysis.
Box 22.8 Identification of miconazole in miconazole nitrate cream by
UV spectrophotometry (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
Cl
N
N
Cl
O
Cl
Cl
Molecular weight:
Acid/base:
pKa:
Log P:
416.1 g/mol
Basic
6.8
6.1
Chemical Analysis of Final Pharmaceutical Products
421
Procedure:
Mix a quantity containing 40 mg of miconazole nitrate with 20 ml of a mixture of 1 vol
of 1 M sulfuric acid and 4 vol of methanol and shake with two 50-ml quantities of
hexane, discarding the organic layers. Make the aqueous phase alkaline with 2 M
ammonia and extract with two 40-ml quantities of chloroform. Combine the chloroform extracts, shake with 5 g of anhydrous sodium sulfate, filter and dilute the filtrate to
100 ml with chloroform. Evaporate 50 ml to dryness and dissolve the residue in 50 ml
of a mixture of 1 vol of 0.1 M hydrochloric acid and 9 vol of methanol. The light
absorption of the resulting solution in the range 230–350 nm exhibits maxima at 264,
272 and 280 nm.
Discussion of the procedure:
UV absorbance
The cream sample is mixed in a two-phase liquid–liquid extraction system, consisting
of a nonpolar phase of hexane and a polar phase of methanol/sulfuric acid. The
hydrophobic excipients are extracted into hexane and discarded, whereas the active
pharmaceutical ingredient (basic substance) is dissolved in the polar phase. Polar
excipients will also remain in the polar phase. However, the polar phase is made
alkaline in a subsequent step; the basic substance miconazole becomes uncharged and
is extracted into chloroform. The chloroform extract is dried with anhydrous sodium
sulfate to remove traces of water, and the extract is evaporated to remove chloroform.
Chloroform is not desired as solvent in the final UV measurement because chloroform
gives strong UV absorption itself. Finally, the residue is dissolved in a mixture of
hydrochloric acid and methanol, this final solution is filled into a cuvette, and the UV
spectrum is recorded between 230 and 350 nm. Absorbance maxima should be
observed at 264, 272 and 280 nm for positive identification of miconazole. The
mixture of methanol and hydrochloric acid is almost UV transparent above
205 nm, and does not contribute to absorption.
264
272
280
200
250
300
350
Wavelength (nm)
400
422
Introduction to Pharmaceutical Chemical Analysis
How much cream should be used?
Cream containing 2% (w/w) miconazole nitrate is to be analyzed. 40 mg (0.040 g)
miconazole nitrate is equivalent to the following amount of ointment:
½ð100%Þ=ð2%Þ 0:040 g 2:0 g
Thus, 2.0 g cream should be used for the identification.
reference substance of the drug substance prepared in a standard solution. If the retention
times are equal, the active pharmaceutical ingredient is positively identified. The major
advantage of HPLC is that dissolved excipients do not have to be removed from the sample,
as long as the HPLC system can separate them from the active pharmaceutical ingredient.
This means that dissolution of the product is the only action that has to be taken before
analysis. Additionally, the same HPLC analysis may also be used for the assay (quantitative
determination) of the active pharmaceutical ingredient. The identification specificity of
HPLC is not comparable to IR, but the specificity obtained by HPLC is acceptable. Several
examples on the use of HPLC for identification are discussed in Boxes 22.9, 22.10, and 22.11.
Box 22.9 Identification of fluoxetine in fluoxetine hydrochloride
capsules by HPLC (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
F
F
F
N
H
Molecular weight:
Acid/base:
pKa:
Log P:
O
309.3 g/mol
Basic
9.8
4.2
Procedure:
For solution (1) dissolve a quantity of the mixed contents of 20 capsules containing the
equivalent of 20 mg fluoxetine in sufficient of the mobile phase to produce 200 ml, filter
and use the filtrate, discarding the first 2 ml. Solution (2) contains 0.011% (w/v) of
fluoxetine hydrochloride (chemical reference substance) in the mobile phase. The
chromatogram obtained with solution (1) shows a peak with the same retention time
as the principal peak in the chromatogram obtained with solution (2).
Chemical Analysis of Final Pharmaceutical Products
423
HPLC conditions:
Column:
Mobile phase:
Detection:
C8 (octyl silyl silica)
33 vol of a solution containing 0.3% (w/v) of glacial acid and 0.64%
(w/v) of sodium pentanesulfonate, adjusted to pH 5.0 with 5 M sodium
hydroxide, and 67 vol of methanol
UV at 227 nm
Discussion of the procedure:
Capsules content corresponding to 20 mg of fluoxetine is added to mobile phase, and
the active pharmaceutical ingredient is dissolved in the mobile phase. Most excipients
are not soluble in the mobile phase, and they are removed by filtration. Since 200 ml
sample solution is produced, but only a very small portion of this is injected in HPLC, the
first 2 ml of the filtrate can be discarded. This is done to saturate the filter with sample
solution, to ensure that no compositional changes occur during filtration of the liquid
injected into the HPLC. The retention time of the active pharmaceutical ingredient in
solution (1) is compared with the retention time for fluoxetine in standard solution (2),
and if they are equal the active pharmaceutical ingredient is identified as fluoxetine. The
chromatographic principle is ion-pair chromatography, where pentanesulfonate is used
as the ion-pairing reagent. With acidic conditions in the mobile phase, fluoxetine is
positively charged and form ion pairs with pentanesulfonate ions. The ion pairs of the
relative nonpolar drug fluoxetine give very strong retention, but this is partly balanced by
a high content of methanol in the mobile phase. Note that both solutions (1) and (2)
contain fluoxetine hydrochloride, but as soon as this salt is dissolved, fluoxetine exists in
its free protonated form in solution. This means that the chloride counter ions do not
affect the retention of fluoxetine.
UV absorbance
API
Solution (1)
Fluoxetine CRS
Solution (2)
Retention time
How much capsule powder should be used?
Capsules containing fluoxetine hydrochloride, each equivalent to 20 mg fluoxetine, is to
be analyzed. 20 capsules are collected, and the total content weight is 3.7520 g. The
average content of each capsule (containing 20 mg of fluoxetine) is:
3:7520 g=20 capsules ¼ 0:1876 g
424
Introduction to Pharmaceutical Chemical Analysis
Capsule content corresponding to 20 mg of fluoxetine is equivalent to one capsule, which
corresponds to 0.1876 g capsule powder. Thus, 0.1876 g capsule powder should be used
for the identification. However, since this is not a quantitative method, exact weighing of
0.1876 g capsule powder is not critical.
Box 22.10 Identification of fentanyl in fentanyl citrate injection by
HPLC (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
N
N
Molecular weight: 336.5 g/mol
Acid/base:
Basic
pKa:
8.8
Log P:
3.8
Procedure:
Solution (1) is the injection, if necessary diluted with the mobile phase, to contain the
equivalent of 0.005% w/v of fentanyl. Solution (2) is a 0.008% w/v solution of fentanyl
citrate (chemical reference substance) in the mobile phase. The chromatogram obtained
with solution (1) shows a peak with the same retention time as the principal peak in the
chromatogram obtained with solution (2).
HPLC conditions:
Column:
end-capped C18 (octadecyl silyl silica)
Mobile phase: 0.3% w/v of potassium dihydrogen orthophosphate in a mixture of 4 vol
of acetonitrile, 40 vol of methanol and 56 vol of water, the solution
adjusted to pH 3.2 with orthophosphoric acid
Detection:
UV at 215 nm
Discussion of the procedure:
In this case, the formulation is very simple, and it can be more or less injected directly
in HPLC. Any excipients will be separated from fentanyl in the HPLC system.
Chemical Analysis of Final Pharmaceutical Products
425
However, if concentrated, the injection has to be diluted in order to avoid over-loading
of the HPLC column and overloading of the UV detector. The retention time of the
active pharmaceutical ingredient in solution (1) is compared with the retention time for
fentanyl in standard solution (2), and if they are equal the active pharmaceutical
ingredient is identified as fentanyl. The chromatographic principle is reversed-phase
chromatography.
How much injection should be used?
Injection containing fentanyl citrate, equivalent to 50 mg/ml fentanyl, is to be analyzed.
This concentration corresponds to:
50 mg=ml ¼ 0:050 mg=ml ¼ 0:005% ðw=vÞ
UV absorbance
API
Solution (1)
Fentanyl CRS
Solution (2)
Retention time
Thus, the injection can be analyzed directly in the HPLC system without further dilution.
TLC, which is discussed earlier in this textbook, can also be used for identification.
However, in modern pharmaceutical laboratories, this technique is less popular than HPLC.
One reason for this is that HPLC is easier to automate, and another reason is that HPLC can
be used for quantitative analysis (assay) at the same time. The latter is not possible with the
same accuracy and precision with TLC. Identification by TLC is similar to identification by
HPLC. Liquid products can be applied more or less directly on the TLC plate, whereas solid
or semi-solid products have to be dissolved in a suitable solvent before application. In
addition to the sample, a reference solution has to be prepared with a chemical reference
substance of the same drug. This reference solution is also applied on the TLC plate, and
after elution, positive identification is obtained if the spot for the active pharmaceutical
ingredient is located at the same distance from the starting line than the spot for the reference
(similar Rf values). The specificity of TLC identifications are not too high, but are
considered to be acceptable. One example on the use of TLC for identification purposes
is illustrated in Box 22.12.
426
Introduction to Pharmaceutical Chemical Analysis
Box 22.11 Identification of beclometasone dipropionate in
beclometasone dipropionate ointment by HPLC (adopted from
British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
O
O
O
H
HO
O
H
H
Cl
H
O
Molecular weight: 408.9 g/mol
Acid/base:
Neutral
pKa:
—
Log P:
2.1
Procedure:
For solution (1) dilute 10 ml of methanol (80%) containing 0.01% w/v of beclometasone
dipropionate to 50 ml with the same solvent. For solution (2) disperse a quantity of the
ointment containing 1 mg of anhydrous beclometasone dipropionate in 25 ml of hot
2,2,4-trimethylpentane, cool and extract the mixture with successive quantities of 20, 10
and 10 ml of methanol (80%), filtering each extract in turn through a small plug of
absorbent cotton previously washed with methanol (80%). Combine the filtrates and
dilute to 50 ml with methanol. The chromatogram obtained with solution (2) shows a
peak with the same retention time as the peak due to beclometasone dipropionate in the
chromatogram obtained with solution (1).
HPLC conditions:
Column:
Mobile phase:
Column temperature:
Detection:
C18 (octadecyl silyl silica)
70 vol of methanol and 30 vol of water
60 C
UV at 238 nm
Discussion of the procedure:
First the ointment sample is dispersed in hot 2,2,4-trimethylpentane to break down the
semi-solid formulation and to give access to the total amount of active pharmaceutical
ingredient. Beclometasone dipropionate is extracted from this very nonpolar phase and
Chemical Analysis of Final Pharmaceutical Products
427
into 80% methanol in water. To ensure an efficient extraction, this is repeated three times
with successive portions of 80% methanol. The extracts are filtered to remove any
particulate matter, and combined prior to the HPLC analysis. The retention time of the
active pharmaceutical ingredient in solution (2) is compared with the retention time for
beclometasone dipropionate in solution (1), and if they are equal the active pharmaceutical ingredient is identified as beclometasone dipropionate. The chromatographic
principle is reversed phase chromatography. Beclometasone dipropionate is a relatively
nonpolar substance, and to avoid too long retention, the content of methanol in the
mobile phase is relatively high.
UV absorbance
Beclometasone dipropionate CRS
Solution (1)
API
Solution (2)
Retention time
How much ointment should be used?
Ointment containing 0.1% (w/w) beclometasone dipropionate is to be analyzed. 1 mg
(0.001 g) beclometasone dipropionate is equivalent to the following amount of ointment:
½ð100%Þ=ð0:1%Þ 0:00100 g 1:0 g
Thus, 1.0 g ointment should be used for the identification. However, since this is not a
quantitative method, exact weighing of 1.0000 g ointment is not critical.
22.4 Assay of the Active Pharmaceutical Ingredient
The assay is intended to check that the content of an active pharmaceutical ingredient is
within a certain tolerance relative to the specified content, that is, to check that the product
contain the correct amount of drug substance. Typically, the analytical result from the assay
should be within a 5.0% tolerance relative to the specified content, but the tolerance may
vary from product to product. The 5.0% is the requirements of the authorities and often
manufactures have stricter limits in order not to risk out of specification problems. In some
cases, particular excipients may also be included in the assay. The major analytical
techniques used for assay are UV spectrophotometry, HPLC, and titration, but also GC
can be used in some cases. In this text, focus is directed towards the three former techniques.
UV spectrophotometry is frequently used for assays, especially for solid formulations
like tablets and for liquid formulations such as injections. The fundamentals of UV have
been discussed in Chapter 7. UV spectrophotometry is always performed on accurately
428
Introduction to Pharmaceutical Chemical Analysis
Box 22.12 Identification of flupentixol decanoate in flupentixol
decanoate injection by TLC (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
O
N
N
F
F
F
S
Molecular weight: 588.8 g/mol
Acid/base:
Basic
Procedure:
Carry out the method for thin-layer chromatography, using silica gel F254 precoated plate
and a mixture of 3 vol of diethylamine and 90 vol of cyclohexane as the mobile phase and
an unsaturated development tank. Apply separately to the plate 5 ml of each of the
following solutions. For solution (1) dilute the injection with ethanol (96%) to contain
0.4% w/v of flupentixol decanoate. Solution (2) contains 0.4% w/v of flupentixol
decanoate dihydrochloride (chemical reference substance) in ethanol (96%). After
removal of the plate, allow it to dry in air, spray with 1% w/v solution of sodium
molybdate in sulfuric acid, heat at 110 C for 20 min and examine in daylight. The
principal spot in the chromatogram obtained with solution (1) is similar in color, position
and size to that in the chromatogram obtained with solution (2).
Discussion of the procedure:
In this procedure, the injection (flupentixol decanoate in a vegetable oil) is first diluted
with ethanol. The injection has to be diluted in order not to overload the TLC plate with
the active pharmaceutical ingredient. Ethanol is used because it is miscible with the
organic injection, and because ethanol is relatively volatile. Volatility is important,
because the solvent has to be evaporated after the sample has been applied onto the TLC
plate. After evaporation, the TLC plate can be eluted with mobile phase to develop the
chromatogram. Solution (2) is used as reference, and if the principal spot for solution (1)
Chemical Analysis of Final Pharmaceutical Products
429
and solution (2) has the same Rf value, the active pharmaceutical ingredient is identified
as flupentixol decanoate. The chromatographic principle is adsorption chromatography.
Pure cyclohexane is a week eluent in adsorption chromatography, but addition of
diethylamine will increase the elution strength and reduce the retention of flupentixol
decanoate because diethylamine interacts strongly with silica as stationary phase.
Solvent front
Flupentixol decanoate
API
Starting line
Solution(1)
Solution(2)
How much injection should be used?
Injection containing 20 mg/ml flupentixol decanoate is to be analyzed. Injection
containing 0.4% w/v (or 4 mg/ml) of flupentixol decanoate is equivalent to the original
injection diluted by a factor of five. Thus, the injection has to be diluted by a factor of five
prior to TLC.
prepared solutions of the active pharmaceutical ingredient, where interfering (UV
absorbing) excipients have been removed prior to the measurement. The latter is
important to make sure that the measured absorbance only arises from the active
pharmaceutical ingredient. Dissolution of the active pharmaceutical ingredient and
removal of UV absorbing excipients recalls on sample preparation. The sample preparation procedures differ from product to product, and some examples on the use of UV for
identification are discussed in Boxes 22.13 and 22.14.
In addition to UV spectrophotometry, HPLC is very popular for assay applications. The
fundamentals of HPLC have been discussed in Chapter 13. The major advantage of HPLC is
that soluble excipients do not have to be removed from the sample, as long as the HPLC
system can separate them from the active pharmaceutical ingredient. This means that
dissolution of the product is the only action that has to be taken before analysis, and
relatively little sample preparation is required. Thus, liquid products can be injected more or
less directly in the HPLC system for analysis, whereas solid or semi-solid products have to
be dissolved in a mobile phase before HPLC. The HPLC system has to be calibrated with a
reference solution of the active pharmaceutical ingredient, containing an exact concentration of chemical reference substance. With this calibration, the relationship between peak
area and concentration is established. Based on this relationship, the peak area measured for
430
Introduction to Pharmaceutical Chemical Analysis
Box 22.13 Assay of paracetamol in paracetamol tablets by UV
spectrophotometry (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
OH
O
N
H
Molecular weight:
Acid/base:
pKa:
Log P:
151.2 g/mol
Acidic (weak)
9.5
0.87
Procedure:
Weigh and powder 20 tablets. Add a quantity of the powder containing 0.15 g of
paracetamol to 50 ml of 0.1 M sodium hydroxide, dilute with 100 ml of water, shake for
15 min and add sufficient water to produce 200 ml. Mix, filter and dilute 10 ml of the
filtrate to 100 ml with water. Add 10 ml of the resulting solution to 10 ml of 0.1 M sodium
hydroxide, dilute to 100 ml with water and measure the absorbance of the resulting
solution at the maximum at 257 nm. Calculate the content of C8H9NO2 taking 715 as the
value of A (1%, 1 cm) at the maximum at 257 nm.
Discussion of the procedure:
The assay is based on powder from 20 tablets to establish a mean value for the content of
active pharmaceutical ingredient. The tablets are powdered to ensure accessibility to the
total amount of active pharmaceutical ingredient. Only a small portion of the powder is
required for the quantitative analysis, namely powder corresponding to 0.15 g of
paracetamol. The required tablet powder is weighed accurately on an analytical balance
and mixed with water containing NaOH. Paracetamol is a weekly acidic compound, and
therefore more soluble in water under alkaline conditions. The mixture is shaken for
15 min to dissolve the total amount of active pharmaceutical ingredient, whereas most
tablet excipients are not dissolved. The later are removed during filtration to avoid any
interference from excipients during the UV measurement. The filtrate is further diluted
before the UV measurement to make sure that the absorbance is in the range 0.2–0.8 as
discussed earlier for UV spectroscopy. Dilution is accomplished in two steps, utilizing
10-ml measuring pipettes (10.00 ml) and a 100-ml volumetric flask (100.0 ml) to ensure
high accuracy. The absorbance is measured at absorption maximum for paracetamol, and
the content of active pharmaceutical ingredient is calculated as shown below. In this
method, the specific absorbance (A; 1%, 1 cm) for paracetamol dissolved in dilute NaOH
has been measured at 257 nm with 1 cm cuvette, and this value is used in Beer’s law to
establish the relationship between measured absorbance and concentration.
Chemical Analysis of Final Pharmaceutical Products
431
How to do the calculations?
Tablets containing 500 mg (0.500 g) paracetamol (specified content) is to be analyzed.
As an example, 20 tablets are collected and the total weight of these 20 tablets is
11.2644 g. The total amount of paracetamol in 20 tablets according to the specification is:
0:500 g 20 tablets ¼ 10:0 g
Tablet powder corresponding to 0.15 g of paracetamol is equivalent to:
½0:15 g=10:0 g 11:2644 g 0:17 g
Thus, approximately 0.17 g tablet powder should be used for the assay. This amount has
to be weighed accurately on an analytical balance because this is a quantitative
measurement, and the exact amount has to be known for the calculations.
As an example, 0.1703 g tablet powder is analyzed by the procedure, resulting in an
absorbance of 0.539 with 1 cm light path. The content of paracetamol per tablet is
calculated as shown below.
The concentration of paracetamol in the final solution measured by UV spectrophotometry is calculated according to Beer’s law:
A=a b ¼ 0:539=½1 cm 715%1 cm1 ¼ 7:54 104 %ðw=vÞ
¼ 7:54 103 mg=ml
The concentration of paracetamol in the initial solution is calculated from the final
solution taking the dilutions into consideration (from 10.00 to 100.0 ml and thereafter
from 10.00 to new 100.0 ml):
7:54 103 mg=ml ½100:0 ml=10:00 ml ½100:0 ml=10:00 ml ¼ 0:754 mg=ml
The total amount of paracetamol in the initial solution is calculated from the concentration and the volume (200.0 ml):
0:754 mg=ml 200:0 ml ¼ 151 mg
This amount of paracetamol was present in 0.1703 g tablet powder.
The average weight of a tablet is:
11:2644 g=20 tablets ¼ 0:5632 g
Therefore, the content of paracetamol in a tablet of average weight is:
½151 mg 0:5632 g=0:1703 g ¼ 499 mg
The deviation from the specified content is equivalent to:
½ð499 mg 500 mgÞ=500 mg 100% ¼ 0:200%
This content is within 5.0% tolerance for the active pharmaceutical ingredient and the
result from the assay is acceptable.
432
Introduction to Pharmaceutical Chemical Analysis
Box 22.14 Assay of doxapram in doxapram hydrochloride injection by
UV spectrophotometry (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
N
N
O
Molecular weight:
Acid/base:
pKa:
Log P:
378.5 g/mol
Basic
7.2
3.2
Procedure:
Dilute a volume containing 0.2 g of doxapram hydrochloride to 250 ml with water.
Measure the absorbance of the resulting solution at the maximum at 258 nm. Calculate
the content of C24H30N2O2,HCl,H2O in the injection from the absorbance of a 0.08% w/v
solution of doxapram hydrochloride BPCRS using the declared content of C24H30N2O2,
HCl,H2O in doxapram hydrochloride BPCRS.
Discussion of the procedure:
Because the injection is a very simple formulation, which contains no other UVabsorbing species, the only sample preparation required is dilution. Dilution is performed to make sure that the measured absorbance is within the range 0.2–0.8 as
discussed under UV spectroscopy (Chapter 7). Dilution is accomplished with a
measuring pipet and a 250 ml volumetric flask (250.0 ml). A 0.08% w/v reference
solution containing chemical reference substance of doxapram hydrochloride is used to
calibrate the UV instrument, to establish the relationship between absorbance and
concentration. The declared content of doxapram hydrochloride in the chemical
reference substance is used to calculate the exact concentration in the reference solution.
How to do the calculations?
Injection containing 20 mg/ml doxapram hydrochloride (specified content) is to be
analyzed. As an example, 10.00 ml of the injection is diluted to 250.0 ml with water, and
the absorbance is measured to 0.766. A reference solution is prepared by dissolving
0.2016 g of doxapram hydrochloride CRS, with a reported content of 99.8%, in 250.0 ml
Chemical Analysis of Final Pharmaceutical Products
433
of water, and the absorbance of this reference solution is measured to 0.753. The content
of doxapram hydrochloride per ml is calculated as shown below.
The exact concentration of doxapram hydrochloride in the reference solution is
calculated taking the reported purity into consideration:
½ðf99:8%g=f100%gÞ 201:6 mg=250 ml ¼ 0:805 mg=ml
This means that 0.805 mg/ml gives an absorbance of 0.753. The absorbance for the
diluted injection was measured to 0.766. The concentration of doxapram hydrochloride
in the diluted injection, which is slightly higher than in the reference solution, is
calculated as follows:
½0:766=0:753 0:805 mg=ml ¼ 0:819 mg=ml
The concentration of doxapram hydrochloride in the original injection is calculated
taking the dilution into consideration (from 10.0 to 250.0 ml):
½250:0 ml=10:0 ml 0:819 mg=ml ¼ 20:5 mg=ml
The deviation from the specified content is equivalent to:
½ð20:5 mg=ml 20:0 mg=mlÞ=20:0 mg=ml 100% ¼ þ2:5%
This content is within 5.0% tolerance for the active pharmaceutical ingredient and the
result from the assay is acceptable.
the final pharmaceutical product can be converted to the exact content of active pharmaceutical ingredient. Several examples on the use of HPLC for assay are discussed in
Boxes 22.15, 22.16, and 22.17.
HPLC and UV spectrophotometry are by far the most popular techniques for assay.
However, titrations are also in use in some cases. The principles of titration have been
discussed in Chapter 5. Both acid–base titrations and redox titrations are in use. A few
examples on the use of titration for assay are discussed in Boxes 22.18 and 22.19.
Box 22.15 Assay of omeprazole in gastro-resistant omeprazole tablets by
HPLC (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
H
N
O
S
O
N
N
O
434
Introduction to Pharmaceutical Chemical Analysis
Molecular weight:
Acid/base:
pKa:
Log P:
345.4 g/mol
zwitterionic
4.8 (basic), 9.3 (acidic)
2.4
Procedure:
Weigh and powder 20 tablets. Carry out the method for liquid chromatography, using the
following solutions.
(1) Disperse a quantity of the powdered tablets containing 24 mg of omeprazole in
150 ml of mobile phase, mix with the aid of ultrasound for 30 min, dilute with
sufficient mobile phase to produce 200 ml, mix, filter, and further dilute 10 vol to
100 vol.
(2) 0.0012% w/v of omeprazole BPCRS in mobile phase.
(3) Mix 10 mg of each of omeprazole BPCRS and omeprazole impurity D BPCRS in
mobile phase and dilute to 100 ml with the same solvent.
HPLC conditions:
Column:
C8 (octyl silyl silica)
Mobile phase: 27 vol of acetonitrile and 73 vol of a 0.14% w/v solution of disodium
hydrogen orthophosphate that has been previously adjusted to pH 7.6
with orthophosphoric acid
Detection:
UV at 280 nm
System suitability:
The test is not valid unless, in the chromatogram obtained with solution (3), the
resolution factor between the peaks due to impurity D and omeprazole is higher than 3.0.
Determination of content:
Calculate the content of C17H19N3O3S in the tablets from the chromatograms obtained
using the declared content of C17H19N3O3S in omeprazole BPCRS.
Discussion of the procedure:
20 tablets are used for the analysis to give a representative sample. The tablets are first
powdered to give access to the active pharmaceutical ingredient. Powder corresponding
to 24 mg omeprazole is required for the analysis. The required tablet powder is weighed
accurately on an analytical balance and is then dispersed in mobile phase to an exact
volume of 200.0 ml. This causes the active pharmaceutical ingredient to be dissolved.
The majority of excipients are not dissolved in the mobile phase, and those are removed
by filtration. The filtrate is further diluted utilizing a 10 ml measuring pipette and a
100 ml volumetric flask to avoid over-loading of the HPLC system. This diluted filtrate is
injected into the HPLC instrument, and the peak area for omeprazole is measured. This
peak area is compared with the peak area of 0.0012% w/v of omeprazole BPCRS in
Chemical Analysis of Final Pharmaceutical Products
435
mobile phase to calculate the content. The declared content of omeprazole in the
chemical reference substance is used to calculate the exact concentration in the reference
solution. The chromatographic principle is reversed-phase chromatography.
How to do the calculations?
Tablets containing 10 mg omeprazole (specified content) is to be analyzed. As an
example, 20 tablets are collected and the total weight is 10.0665 g. The total amount of
omeprazole in 20 tablets according to the specification is:
10:0 mg 20 tablets ¼ 200 mg
Tablet powder corresponding to 24 mg of omeprazole is equivalent to:
ð24 mg=200 mgÞ 10:0665 g 1:2 g
Thus, approximately 1.2 g tablet powder should be used for the assay. The tablet powder
has to be weighed accurately on an analytical balance because this is a quantitative
measurement.
As an example, 1.2097 g tablet powder is analyzed by the procedure, resulting in a
peak area of 109 766 for the active pharmaceutical ingredient. The reference solution (2)
is prepared from omeprazole BPCRS of reported purity 99.9%. 0.1213 g omeprazole
BPCRS is dissolved in mobile phase and the volume is adjusted to 100.0 ml. 10.00 ml of
this solution is diluted to 100.0 ml with mobile phase, and 10.00 ml of this solution is
diluted further to 100.0 ml with mobile phase. The peak area of solution (2) for
omeprazole is 108 871.
Area =
109 766
API
UV absorbance
Solution (1)
Omeprazole CRS
Area =
108 871
Solution (2)
Omeprazole
impurity D
Omeprazole CRS
Solution (3)
Retention time
The content of omeprazole per tablet is calculated as shown below.
436
Introduction to Pharmaceutical Chemical Analysis
The concentration of omeprazole in reference solution (2) is calculated based on the
reported purity and taking the dilutions into consideration (from 10.00 to 100.0 ml and
thereafter from 10.00 to new 100.0 ml):
½121:3 mg=100:0 ml ½99:9%=100:0% ½10:00 ml=100:0 ml
½10:00 ml=100:0 ml ¼ 0:0121 mg=ml
This means that 0.0121 mg/ml gives a peak area of 108 871. The peak area for the final
sample solution was measured to 109 766. The concentration of omeprazole in the final
solution (1), which is slightly higher than in the reference solution (2), is calculated as
follows:
½109 766=108 871 0:0121 mg=ml ¼ 0:0122 mg=ml
The original amount of omeprazole in the sample is calculated from the concentration in
the final solution (1), taking the dilution into consideration (from 10.00 to 100.0 ml) and
the total volume of the original volumetric flask (200.0 ml):
0:0122 mg=ml ½100:0 ml=10:0 ml 200:0 ml ¼ 24:4 mg
This amount of omeprazole is present in 1.2097 g tablet powder.
The average weight of a tablet is:
10:0665 g=20 tablets ¼ 0:5033 g
Therefore, the content of omeprazole in a tablet of average weight is:
½24:4 mg 0:5033 g=1:2097 g ¼ 10:2 mg
The deviation from the specified content is equivalent to:
½ð10:2 mg 10:0 mgÞ=10:0 mg 100% ¼ þ2:00%
This content is within 5.0% tolerance for the active pharmaceutical ingredient.
We also have to check the system suitability test. Calculate the resolution yourself
from the chromatogram above, and check that the value is higher than 3.0.
Box 22.16 Assay of fentanyl in fentanyl citrate injection by HPLC
(adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
N
N
Chemical Analysis of Final Pharmaceutical Products
437
Molecular weight: 336.5 g/mol
Acid/base:
Basic
pKa:
8.8
Log P:
3.8
Procedure:
Carry out the method for liquid chromatography, using the following solutions. Solution
(1) is the injection, diluted if necessary with the mobile phase, to contain the equivalent
of 0.005% w/v of fentanyl. Solution (2) is a 0.008% w/v solution of fentanyl citrate
BPCRS in the mobile phase.
HPLC conditions:
Column:
Mobile phase:
Detection:
C18 (octadecyl silyl silica)
0.3% w/v potassium dihydrogen orthophosphate in a mixture of 4 vol of
acetonitrile, 40 vol of methanol and 56 vol of water, adjusted to pH 3.2
with orthophosphoric acid
UV at 215 nm
Determination of content:
Calculate the content of C22H28N2O in the injection using the declared content of
C22H28N2O in fentanyl citrate BPCRS.
Discussion of the procedure:
The injection is diluted to avoid overloading of the HPLC system, and then injected
directly without any further sample preparation, and the peak area for the active
pharmaceutical ingredient is measured. This peak area is compared with the peak
area of 0.008% w/v solution of fentanyl citrate BPCRS in mobile phase to calculate the
content of fentanyl. The declared content of fentanyl citrate in the chemical reference
substance is used to calculate the exact concentration in the reference solution. The
chromatographic principle is reversed-phase chromatography.
How to do the calculations?
Injection containing fentanyl citrate, corresponding to 50 mg/ml of fentanyl is to be
analyzed. 50 mg/ml corresponds to 0.050 mg/ml ¼ 0.0050% w/v, and the injection can be
injected directly into the HPLC system according to the procedure. As an example, direct
injection of the injection provides a principal peak with an area of 99 753. The reference
solution (2) is prepared from fentanyl citrate BPCRS of reported purity 99.7%. Fentanyl
citrate BPCRS of 0.0810 g (81.0 mg) is dissolved in mobile phase and the volume is
adjusted to 100.0 ml. 10.00 ml of this solution is diluted to 100.0 ml with mobile phase.
The peak area of solution (2) for fentanyl is 100 763.
Introduction to Pharmaceutical Chemical Analysis
UV absorbance
438
Area =
99 753
API
Solution (1)
Fentanyl CRS
Area =
100 763
Solution (2)
Retention time
The content of fentanyl per ml is calculated as shown below.
The concentration of fentanyl citrate in reference solution (2) is calculated based
on the reported purity and taking the dilutions into consideration (from 10.00 to
100.0 ml):
½81:0 mg=100:0 ml ½99:7%=100:0% ½10:00 ml=100:0 ml ¼ 0:0808 mg=ml
This means that 0.0808 mg/ml fentanyl citrate gives a peak area of 100 763. The peak
area for the final sample solution was measured to 99 753. The concentration of fentanyl
citrate in the final solution (1), which is slightly lower than in the reference solution (2), is
calculated as follows:
½99 753=100 763 0:0808 mg=ml ¼ 0:0799 mg=ml
This is the concentration of fentanyl citrate. To calculate the concentration of fentanyl,
one has to take the molar weights of fentanyl (336.5 g/mol) and for fentanyl citrate
(528.6 g/mol) into consideration. The content of fentanyl in the injection is calculated
as follows:
0:0799 mg=ml ½336:5 g=mol=528:6 g=mol ¼ 0:0509 mg=ml ¼ 50:9 mg=ml
The deviation from the specified content is equivalent to:
½ð50:9 mg=ml 50:0 mg=mlÞ=50:0 mg=ml 100% ¼ þ1:80%
This content is within 5.0% tolerance for the active pharmaceutical ingredient.
Chemical Analysis of Final Pharmaceutical Products
439
Box 22.17 Assay of hydrocortisone in hydrocortisone ointment by
HPLC (modified from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
OH
OH
HO
H
H
H
O
Molecular weight: 362.5 g/mol
Acid/base:
neutral
pKa:
—
Log P:
1.2
Procedure:
Carry out the method for liquid chromatography using the following solutions (for
ointments containing more than 0.5% w/w of hydrocortisone). For solution (1) dissolve
25 mg of hydrocortisone BPCRS in 45 ml of methanol, add 5 ml of a 0.5% w/v solution of
betamethasone (internal standard) in methanol and add sufficient water to produce
100 ml. For solution (2), disperse a quantity of ointment containing 25 mg of hydrocortisone in 100 ml of hot hexane, cool and extract with 20 ml of a solution prepared by
mixing 3 vol of methanol with 1 vol of a 15% w/v solution of sodium chloride. Repeat the
extraction using further two 10 ml quantities of the methanolic sodium chloride solution.
To the combined extracts add 5 ml of a 0.5% w/v solution of betamethasone (internal
standard) in methanol and sufficient water to produce 100 ml, mix and filter through a
glass microfiber filter (Whatman GF/C is suitable).
HPLC conditions:
Column:
Mobile phase:
Detection:
C18 (octadecyl silyl silica)
methanol mixed with water (50:50 v/v)
UV at 240 nm
Determination of content:
Calculate the content of C21H30O5 in the preparation being examined using the declared
content of C21H30O5 in hydrocortisone BPCRS.
440
Introduction to Pharmaceutical Chemical Analysis
Discussion of the procedure:
First the ointment is dispersed in hot hexane. Hot hexane disintegrates the semi-solid
formulation, and the total amount of active pharmaceutical ingredient becomes
accessible. This hexane dispersion is then extracted (liquid–liquid extraction) with a
mixture of methanol and water. This mixture forms a two-phase system with hexane.
The methanol/water phase contains large amounts of sodium chloride in order to
increase the polarity of the extracting phase. Hydrocortisone is a relatively hydrophilic compound, and is extracted into the methanol/water phase. This extraction is
repeated with two more portions of methanol/water to ensure a quantitative transfer
of hydrocortisone from the ointment. The three extracts are combined, and a small
amount of betamethasone is added as internal standard. Hydrocortisone is measured
relative to the internal standard, to improve the precision and the accuracy of the
method. The combined extracts are diluted to 100.0 ml in a volumetric flask, and a
small portion of this is filtered prior to HPLC. Filtration is done to remove any
particulate matter, which can plug the HPLC column. The peak area for solution
(1) is compared with the peak area of reference solution (2) to calculate the
content of active pharmaceutical ingredient. The declared content of hydrocortisone
in the chemical reference substance is used to calculate the exact concentration
in the reference solution. The chromatographic principle is reversed-phase
chromatography.
How to do the calculations?
Ointment containing 1% w/w hydrocortisone is to be analyzed. As an example, the
internal standard solution is prepared by dissolution of 0.1249 g betamethasone in
25.0 ml methanol.
Solution (1) is prepared as specified by dissolution of 25.2 mg hydrocortisone BPCRS
with a specified content of 99.8% w/w. The concentration of hydrocortisone in solution
(1) is calculated based on the reported purity:
½25:2 mg=100:0 ml ½99:8%=100:0% ¼ 0:252 mg=ml
As an example, 2.4896 g ointment was prepared as specified for solution (2). For
solution (1), the peak areas measured for hydrocortisone and for the internal standard
were 87 635 and 56 933, respectively. For solution (2), the peak areas measured for
hydrocortisone and for the internal standard were 86 701 and 57 255, respectively.
The content of hydrocortisone in the ointment is calculated as follows. For solution
(1), which is a reference solution, a concentration of 0.252 mg/ml provided a peak area
ratio of
87 635=56 933 ¼ 1:539
Chemical Analysis of Final Pharmaceutical Products
UV absorbance
Hydrocortisone
Area =
87 635
API
Area =
86 701
441
Internal standard
(betamethasone)
Area =
56 933
Solution (1)
Internal standard
Area =
57 255
Solution (2)
Retention time
For solution (2), which is the sample solution, the following peak area ratio was
obtained:
86 701=57 255 ¼ 1:514
This means that the concentration in the sample solution was slightly lower than in the
reference solution. The concentration in the sample solution is calculated as follows:
½1:514=1:539 0:252 mg=ml ¼ 0:248 mg=ml
The volume of the final solution was 100.0 ml, and the total amount of hydrocortisone in
this solution was:
0:248 mg=ml 100:0 ml ¼ 24:8 mg
This amount of hydrocortisone was included in 2.4896 g (2489.6 mg) ointment, and the
concentration within the formulation is calculated as follows:
½24:8 mg=2489:6 mg 100% w=w ¼ 0:996% w=w
The deviation from the specified content is equivalent to:
½ð0:996% w=w 1:00% w=wÞ=1:00% w=w 100% ¼ 0:386%
This content is within 5.0% tolerance for the active pharmaceutical ingredient.
442
Introduction to Pharmaceutical Chemical Analysis
Box 22.18 Assay of Fe2þ in ferrous fumarate tablets by titration
(adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
Fe2+
O-
O
O
Molecular weight: 169.9 g/mol
Acid/base:
—
pKa:
—
Log P:
—
Procedure:
Weigh and powder 20 tablets. Dissolve a quantity of the powder containing 0.3 g of
ferrous fumarate in 7.5 ml of 1 M sulfuric acid with gentle heating. Cool, add 25 ml of
water and titrate immediately with 0.1 M ammonium cerium(IV) sulfate VS using
ferroin solution as indicator. Each 1.0 ml of 0.1 M ammonium cerium(IV) sulfate VS is
equivalent to 5.585 mg of Fe(II).
Discussion of the procedure:
20 tablets are collected to have a representative sample, and the tablets are powdered to
make the total amount of active pharmaceutical ingredient accessible to the assay. Only a
minor fraction is required to do the titration, namely powder corresponding to 0.3 g of
ferrous fumarate. This amount of powder is dissolved in sulfuric acid, and heating is
required to promote dissolution. Water is added to increase the volume of the sample
solution, and titration is performed immediately in order to make sure that Fe2þ is not
oxidized by air to Fe3þ prior to the titration. The titration is a redox titration according to
the following reaction:
Fe2þ þ Ce4þ ! Fe3þ þ Ce3þ
Ferroin is a color-indicator that detects the endpoint for the titration; the point where the
total amount of Fe2þ from the sample has reacted with Ce4þ. The consumption of Ce4þ is
measured and used to calculate the exact content of Fe2þ.VS behind ammonium cerium
(IV) sulfate stands for volumetric solution and indicate that this is a solution for titration
with exactly known concentration.
How to do the calculations?
Tablets containing 50 mg (0.050 g) Fe2þ is to be analyzed. As an example, 20 tablets are
collected and the total weight is 12.3673 g. The total amount of Fe2þ in 20 tablets
according to the specification is:
0:050 g 20 tablets ¼ 1:0 g
Chemical Analysis of Final Pharmaceutical Products
443
Tablet powder corresponding to 0.3 g of ferrous fumarate (molar weight of 169.9 g/mol)
is equivalent to the following amount of Fe2þ (molar weight of 55.847 g/mol):
0:3 g ½55:847 g=mol=169:9 g=mol ¼ 0:1 g
0.1 g Fe2þ is equivalent to the following amount of tablet powder:
½0:1 g=1:0 g 12:3673 g 1:2 g
As an example, 1.2109 g tablet powder is weighed on the analytical balance for the assay
and titrated with 17.62 ml of 0.1006 M ammonium cerium(IV) sulfate. The content of
Fe2þ per tablet is calculated as shown below.
The consumption of ammonium cerium(IV) sulfate is calculated from the titration
volume and the molarity:
½17:62 ml=1000 ml 0:1006 M ¼ 1:772 103 mol
This number of moles was equivalent to the number of moles of Fe2þ in 1.2109 g tablet
powder, which corresponded to the following weight found by multiplication with the
molecular weight for Fe2þ (55.847 g/mol):
1:773 103 mol 55:847 g=mol ¼ 0:0990 g ¼ 98:99 mg
This amount of Fe2þ was present in 1.2109 g tablet powder.
The average weight of a tablet is:
12:3673 g=20 tablets ¼ 0:6184 g
Therefore, the content of Fe2þ in a tablet of average weight is:
½98:99 mg 0:6184 g=1:2109 g ¼ 50:55 mg
The deviation from the specified content is equivalent to:
½ð50:55 mg per tablet 50 mg per tabletÞ=50 mg per tablet 100% ¼ þ1:1%
This content is within 5.0% tolerance for the active pharmaceutical ingredient.
444
Introduction to Pharmaceutical Chemical Analysis
Box 22.19 Assay of diphenhydramine in diphenhydramine
hydrochloride oral solution by titration (adopted from British
Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
N
O
Molecular weight: 255.3 g/mol
Acid/base:
basic
pKa:
8.9
Log P:
3.7
Procedure:
Acidify a quantity containing 0.1 g diphenhydramine hydrochloride with 2 M hydrochloric acid, shake with three 20-ml quantities of ether, discard the ether, make the
aqueous solution alkaline with 5 M sodium hydroxide and extract with successive 15-ml
quantities of ether until extraction is complete. Wash the combined ether extracts with
two 5-ml quantities of water, extract the combined washings with 15 ml of ether and
evaporate the combined ether extracts to dryness. Dissolve the residue in 15 ml of 0.05 M
sulfuric acid VS and titrate the excess of acid with 0.1 M sodium hydroxide VS using
methyl red solution as indicator. Each 1 ml of 0.05 M sulfuric acid VS is equivalent to
29.18 mg of C17H21NO,HCl.
Discussion of the procedure:
First, the oral solution (aqueous) is acidified to ensure high solubility of the basic
active pharmaceutical ingredient. Then, the oral solution is extracted with ether. In
this extraction, important excipients are extracted into ether, whereas the active
pharmaceutical ingredient remains in the aqueous phase. Removal of excipients is
important to avoid that they interfere during the titration. Then, the aqueous oral
solution is made alkaline with sodium hydroxide to reduce the solubility of the active
pharmaceutical ingredient, and this is extracted into ether. The ether extracts are
washed with water to remove further excipients, and the water washings are extracted
with ether again to make sure that no diphenhydramine is lost during this quantitative
procedure. Finally, the ether extracts are combined, and the ether is evaporated. The
residue, which is the active pharmaceutical ingredient (diphenhydramine), is
Chemical Analysis of Final Pharmaceutical Products
445
dissolved in an exact amount of sulfuric acid in excess. Sulfuric acid reacts with
diphenhydramine by protonation:
2Diphenhydramine þ H2 SO4 ! 2DiphenhydramineHþ þ SO4 2
The excess amount of sulfuric acid, which has not reacted with the active pharmaceutical
ingredient, is then titrated with sodium hydroxide using methyl red as indicator
according to the following reaction:
H2 SO4 þ 2OH ! SO4 2 þ 2H2 O
Based on the difference between the total amount of sulfuric acid added and the excess
titrated by sodium hydroxide, the amount of diphenhydramine can be calculated.
How to do the calculations?
Oral solution containing 2.5 mg/ml of diphenhydramine is to be analyzed. 0.1 g (100 mg)
diphenhydramine is equivalent to:
100 mg=ð2:5 mg=mlÞ ¼ 40 ml
As an example, 40.00 ml oral solution is collected by measuring pipet and analyzed
according to the procedure. (Note that oral solutions may be quite viscous, and therefore
can be difficult to pipette. In such cases an exact amount of sample is weighed on an
analytical balance, and the exact volume is calculated from the exact mass and the
density of the product.) The residue is dissolved in 15.00 ml of 0.0493 M sulfuric acid,
and excess sulfuric acid is titrated with 10.65 ml of 0.1026 M NaOH. The content of
diphenhydramine/ml is calculated as shown below.
The total amount of H2SO4 added to the residue is calculated from the volume and
from the molarity:
½15:00 ml=1000 ml 0:0493 M ¼ 7:40 104 mol
The amount of sodium hydroxide used during the final titration is calculated from the
volume and from the molarity:
½10:65 ml=1000 ml 0:1026 M ¼ 1:093 103 mol
This amount of sodium hydroxide reacted 2 : 1 with excess sulfuric acid, and excess
sulfuric acid is calculated as:
0:5 1:093 103 mol ¼ 5:464 104 mol
The number of moles of sulfuric acid reacted with diphenhydramine can be calculated as
follows:
7:40 104 mol 5:464 104 mol ¼ 1:93 104 mol
446
Introduction to Pharmaceutical Chemical Analysis
Sulfuric acid reacted 1 : 2 with diphenhydramine, and therefore the amount of diphenhydramine in the sample is calculated as follows:
2 1:93 104 mol ¼ 3:86 104 mol
This corresponded to the following weight of diphenhydramine found by multiplication
with the molecular weight for diphenhydramine (255.3 g/mol):
3:86 104 mol 255:3 g=mol ¼ 0:0986 g ¼ 98:6 mg
This weight was present in 40.00 ml of oral solution, and the concentration in the oral
solution is calculated as follows:
98:6 mg=40:00 ml ¼ 2:46 mg=ml
The deviation from the specified content is equivalent to:
½ð2:46 mg=ml 2:50 mg=mlÞ=2:50 mg=ml 100% ¼ 1:60%
This content is within 5.0% tolerance for the active pharmaceutical ingredient.
22.5 Chemical Tests for Final Pharmaceutical Products
In addition to identification (Section 22.3) and assay (Section 22.4) of the active pharmaceutical ingredient, several tests are performed to check the quality of the final pharmaceutical product. This section briefly discusses some of these tests based on chemical methods.
It is important to emphasize that this is a selection of some of the common tests and not a
complete overview. It is also important to emphasize that final pharmaceutical products are
subjected to pharmaceutical and microbiological testing as well, but this is not covered by
this textbook.
22.5.1
Test for Related Substances
For many final pharmaceutical products, a test for related substances is performed as a part
of the chemical quality control. This test is intended to check that the content of impurities
structurally related to the active pharmaceutical ingredient (related substances) is at a very
low level. This is important as related substances can affect the efficacy of the final
pharmaceutical product. Especially, degradation products from the active pharmaceutical
ingredient are important. Test for related substances can be performed as a part of the
release control for a production batch, or can be performed on products stored for different
periods of time and under different conditions to establish the shelf life for a pharmaceutical product. The latter is called stability testing and is typically performed during
pharmaceutical development. The test for related substances is normally performed by
HPLC. The pharmaceutical product is typically dissolved or dispersed in a mobile phase to
dissolve the active pharmaceutical ingredient and the related substances, and this is
Chemical Analysis of Final Pharmaceutical Products
447
analyzed after filtration by HPLC to detect related substances with retention times
different from the active pharmaceutical ingredient. Typically, any peak for a related
substance is compared with a reference solution serving as the limit. If any peak for related
substances is greater than the peak for the reference, the level of related substances is too
high in the final pharmaceutical product. One example of a test for related substances is
shown in Box 22.20 for paracetamol tablets.
Box 22.20 Related substances in paracetamol tablets by HPLC (adopted
from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
OH
O
N
H
Molecular weight: 151.2 g/mol
Acid/base:
Acidic (weak)
pKa:
9.5
Log P:
0.87
Procedure:
Carry out the method for liquid chromatography, using the following solutions. Prepare
the solutions immediately before use and protect from light. For solution (1) disperse a
quantity of powdered tablets containing 0.2 g of Paracetamol in 8 ml of the mobile phase
with the aid of ultrasound, add sufficient mobile phase to produce 10 ml, mix well and
filter. For solution (2) dilute 1 vol of solution (1) to 20 vol with mobile phase and dilute
1 vol of this solution to 20 vol with mobile phase. Solution (3) contains 0.002% w/v each
of 4-aminophenol and paracetamol BPCRS in the mobile phase. Solution (4) contains
0.00002% w/v of 40 -chloroacetanilide in the mobile phase.
The chromatographic procedure may be carried out using: (a) a stainless steel column
(25 cm 4.6 mm) packed with octylsilyl silica gel for chromatography (5 mm; Zorbax
Rx C8 is suitable), (b) as the mobile phase with a flow rate of 1.5 ml/min, at a temperature
of 35 C, a mixture of 250 vol of methanol containing 1.15 g of a 40% v/v solution of
tetrabutylammonium hydroxide with 375 vol of 0.05 M disodium hydrogen orthophosphate and 375 vol of 0.05 M sodium dihydrogen orthophosphate and (c) a detection
wavelength of 245 nm.
The test is not valid unless, in the chromatogram obtained for solution (3), the
resolution factor between the two principal peaks is at least 4.0.
Inject solution (1) and allow the chromatography to proceed for 12 times the
retention time of the principal peak. In the chromatogram obtained with solution (1) the
448
Introduction to Pharmaceutical Chemical Analysis
area of any peak corresponding to 4-aminophenol is not greater than the area of the
corresponding peak in solution (3) (0.1%), the area of any peak corresponding to 40 chloroacetanilide is not greater than the area of the principal peak in solution (4)
(10 ppm) and no other impurity is greater than the area of the principal peak obtained
with solution (2) (0.25%).
Discussion of the procedure:
In solution (1), tablet powder is dispersed in mobile phase. Paracetamol and eventual
related substances are dissolved in the mobile phase, whereas the majority of excipients
are not dissolved. The latter are therefore removed by filtration. Paracetamol is present at
the 20 mg/ml level (2.0% w/v) in Solution (1). This is a very high concentration, but is
required to be able to detect trace levels of related substances. Solution (1) is injected into
the HPLC system, and any peaks apart from the peak for paracetamol are observed.
Among the related substances, 4-aminophenol is of special interest. The retention time
for 4-aminophenol is obtained from solution (3), which contains 0.002% (w/v) of 4aminophenol. If any peak is observed at this retention time for solution (1), 4aminophenol is present in the tablets. In such a case, the area of the peak for 4aminophenol in solution (1) should be less than the area for 4-aminophenol in solution
(3). This corresponds to the following limit calculated relative to the amount of
paracetamol in the tablets:
½0:002%ðw=vÞ=2:0%ðw=vÞ 100% w=w ¼ 0:1% w=w
In addition, 40 -chloroacetanilide is also of special interest. The retention time for this
related substance is obtained from solution (4), and if 40 -chloroacetanilide is present in
solution (1), the peak area should be less in solution (1) as compared to solution (4). This
corresponds to the following limit calculated relative to the amount of paracetamol in
the tablets:
½0:00002%ðw=vÞ=2:0%ðw=vÞ 106 ¼ 10 ppm
Any other impurities should not be observed with peak areas exceeding the principal
peak in solution (2). Solution (2) contains paracetamol at the following concentration
level, taking the 1:20 dilution followed by 1:20 dilution from solution (1) into
consideration:
2:0%ðw=vÞ ½1=20 ½1=20 ¼ 0:0050% w=v
Assuming that other impurities have the same specific absorbance at 245 nm, this
corresponds to the following limit calculated relative to the amount of paracetamol in the
tablets:
½0:0050%ðw=vÞ=2:0%ðw=vÞ 100%ðw=wÞ ¼ 0:25%ðw=wÞ
Chemical Analysis of Final Pharmaceutical Products
449
A chromatogram for a test for related substances in paracetamol tablets are shown below.
Paracetamol
UV absorbance
N-(3-Chloro-4-hydroxy phenyl)
acetamide
4-Hydroxy
propanamide
4-Hydroxy acetophenone
oxime
Retention time
4-Aminophenol
22.5.2
4-Nitro phenol
4'-Chloroacetamide
Uniformity of Content
The test Uniformity of content is intended to check that the amount of active pharmaceutical
ingredient does not vary too much from dose unit to dose unit. This test is typically used for
final pharmaceutical products containing 2 mg or less of active pharmaceutical ingredient
per dosage unit, or in other cases where the dose to dose variation is critical. In this test,
typically 10 individual dosage units are collected randomly, like 10 individual tablets, and
each dosage unit is analyzed individually to determine the amount of active pharmaceutical
ingredient in each unit. The results for the 10 individual dosage units should be within a
certain specified limit in order for the product to comply with the test. The quantitative
method used for Uniformity of content is often the same method as used for the assay, the
only difference being that the assay gives an average value for the content. Box 22.21
illustrates an example of the test Uniformity of content for phenindione.
Box 22.21 Uniformity of content for phenindione tablets by UV
spectrophotometry (adopted from British Pharmacopoeia 2009)
Chemical properties of the active pharmaceutical ingredient:
O
O
450
Introduction to Pharmaceutical Chemical Analysis
Molecular weight: 222.2 g/mol
Acid/base:
Neutral
pKa:
—
Log P:
3.0
Procedure:
Tablets containing 50 mg or less of Phenindione comply with the requirements stated
under Tablets using the following method of analysis. Place one tablet in 50 ml of 0.1 M
sodium hydroxide, dissolve the tablet completely by shaking gently, add a further 100 ml
of 0.1 M sodium hydroxide and shake for 1 h. Dilute to 250 ml with 0.1 M sodium
hydroxide, filter, and dilute a portion of the filtrate with sufficient 0.1 M sodium
hydroxide to produce a solution containing 4 mg/ml of Phenindione. Measure the
absorbance of the resulting solution at the maximum at 278 nm. Calculate the content
of C15H10O2 taking 1310 as the value of A (1%, 1 cm) at the maximum at 278 nm.
Perform this procedure on 10 individual and randomly selected tablets.
Criteria:
The preparation complies with the test if each individual content is between 85 and 115%
of the average content. The preparation fails to comply with the test if more than one
individual content is outside these limits or if one individual content is outside the limits
of 75–125% of the average content.
Discussion of the procedure:
This quantitative method is performed separately on 10 individual and randomly
collected tablets, and the 10 results are checked according to the criteria. The quantitative
analysis is a relatively simple UV method in this case, very similar to the example in
Box 22.13.
How to do the calculations?
As an example, the following results (all in mg per tablet) are obtained for tablets
specified to contain 10 mg phenindione (in increasing order of determined content):
10:1
10:1
10:3
10:4 10:4
10:7 10:8
11:1 11:2
11:4
The average content is calculated to be 10.65 mg per tablet. 85% and 115% of this value
corresponds to:
½ð85%Þ=ð100%Þ 10:65 mg per tablet ¼ 9:05 mg per tablet
½ð115%Þ=ð100%Þ 10:65 mg per tablet ¼ 12:25 mg per tablet
Clearly, all 10 tablets are within 85–115% of the average content, and the results comply
with the criteria.
Chemical Analysis of Final Pharmaceutical Products
22.5.3
451
Dissolution
Dissolution is a test intended to measure the rate of release of active pharmaceutical
ingredients into solution from the final pharmaceutical product. Dissolution testing is
performed for tablet or capsule formulations. Information about the dissolution rate is
important because it may have an important effect on the therapeutic efficiency of the
product. Basically, the dissolution test involves placing the tablet or capsule inside a
stainless steel wire basket, which is rotated at a fixed speed while immersed in a dissolution
medium. The steel wire basket and the dissolution medium are contained in a wide-mouthed
cylindrical vessel. Samples of the dissolution medium are collected at specified times,
filtered, and assayed typically by UV spectroscopy. In this way, the build up of active
pharmaceutical ingredient in the dissolution medium can be measured versus time to check
that the drug is released properly from the formulation.
23
Analysis of Drugs in Biological Fluids
This chapter focuses on analytical methods for the determination of drugs in biological
fluids and on the application of bioanalytical methods in drug development, therapeutic
drug monitoring, toxicological analysis and doping control analysis. Knowledge of the
chemical composition of biological matrices such as blood, serum, plasma and urine is
crucial for bioanalytical work and is discussed in detail. The various steps of a bioanalytical
method include sampling, sample preparation, separation, detection, identification, calibration and quantification. These steps are discussed on a general basis and highlighted by
examples discussing various sample preparation procedures, quantitative determinations
and identification of unknown components.
23.1 Introduction
Bioanalysis is a subdiscipline of pharmaceutical analysis covering the analysis of drug
substances and their metabolites in biological fluids. It is a fundamental part of drug
development programs in the pharmaceutical industry, in therapeutic drug monitoring and
in many other areas, such as forensic science, forensic toxicology and doping testing
in sports.
23.1.1
Drug Development
Figure 23.1 shows a schematic view of the drug development process. Drug development
begins with the identification of a medical need and hypotheses on how therapy can be
improved. Drug discovery is the identification of drug candidates based on combinatorial
chemistry, high-throughput screening, genomics and ADME (absorption, distribution,
metabolism, elimination). By combinatorial chemistry a great number of compounds
are synthesized fast and these are tested for pharmacological activity and potency in
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
454
Introduction to Pharmaceutical Chemical Analysis
high throughput screening systems (HTS). The HTS systems simulate the interaction of the
compounds with a specific biological receptor or target. Once a lead compound is found a
narrow range of similar compounds is synthesized and screened to improve the activity
toward the specific target. Other studies investigate the ADME profile of drug candidates by
analyzing samples collected at different time points from dosed laboratory animals (in vivo
testing) and tissue cultures (in vitro testing). Drug candidates passing the discovery phase
are subjected to toxicity testing and further metabolism and pharmacological studies in the
preclinical phase. Both in vivo and in vitro tests are conducted and various animal species
are used to prove the pharmacokinetic profile of the candidate. The detailed information
about the candidate form the basis for further pharmaceutical research on the synthesis and
raw materials, dosage forms, quality control and stability testing.
The clinical development phase can begin when a regulatory body has judged a drug
candidate to be effective and to appear safe in healthy volunteers. In phase I the goal is to
establish a safe and efficient dosage regimen and to assess pharmacokinetics. Blood samples
are collected and analyzed from a small group of healthy volunteers (20–80). The data
obtained form the basis for developing controlled phase II studies. The goal of phase II
studies is to demonstrate a positive benefit/risk balance in a larger group of patients
(200–800) and to further study pharmacokinetics. Monitoring of efficacy and monitoring of
possible side effects are essential. Phase II studies can take up to two years to fulfil. At the
end of phase II a report is submitted to the regulatory body and conditions for phase III
studies are discussed. Additional information supporting the claims for a new drug is
provided. Phase III begins when evidence for the efficacy of the drug candidate and
supporting data has demonstrated a favorable outcome to the regulatory body. The phase III
studies are large-scale efficacy studies with focus on the effectiveness and safety of the drug
candidate in a large group of patients. In most cases the drug candidate is compared with
another drug used to treat the same condition. Phase III studies can last two or three years or
more and 3000–5000 patients can be involved. Carcinogenetic tests, toxicology tests and
metabolic studies in laboratory animals are conducted in parallel. The cumulative data form
the basis for filing a new drug application to the regulatory body and for future plans for
manufacturing and marketing. The regulatory body thoroughly evaluates the documentation provided before a market approval eventually can be authorized and the drug product
can be legally marketed. As shown in Figure 23.1 the time required from drugs discovery to
product launch is up to 12 years. Phase IV studies are studies conducted after product launch
to demonstrate long-term effects, new claims, to expand on approved claims, to examine
possible drug–drug interactions and to further assess pharmacokinetics. Several thousand
patients participate in phase IV studies.
Drug
discovery
Clinical development
Preclinical
development
Phase I
Phase II
Phase III
Regulatory
approval
Product
launch
6 to 12 years
Figure 23.1 Schematic diagram of the drug development process
Analysis of Drugs in Biological Fluids
455
The need for the industry to develop new drugs faster and to reduce costs in the
development process is a major driving force for inventions in new analytical instrumentation and new technologies. A combination of robotics, liquid handling workstations and new
formats for sample preparation now allows high-speed analysis. New principles for
chromatographic separation and new instrumentation, particularly in LC-MS, have greatly
reduced time and costs in the identification of unknown substances and in the quantification
of known substances in biological fluids. These innovations are also being implemented in
laboratories conducting therapeutic drug monitoring, doping control in sports and in
forensic and toxicological science.
23.1.2
Therapeutic Drug Monitoring
Therapeutic drug monitoring (TDM) refers to the individualization of dosage by maintaining serum or plasma drug concentrations within a target range to optimize efficacy and to
reduce the risk of adverse side effects. The target range of a drug is also called the
therapeutic range or the therapeutic window and is the concentration range between the
lowest drug concentration that has a positive effect and the concentration that gives more
adverse effects than positive effects. Variability in the dose–response relationship between
individual patients is due to pharmacokinetic and pharmacodynamic variability as shown in
Figure 23.2.
Pharmacodynamic variability arises from variations in drug concentrations at the
receptor and in variations in the drug–receptor interaction. Pharmacokinetic variability
is due to variations in the dose to plasma concentration relationship. Major sources of
pharmacokinetic variability are age, physiology, disease, compliance and genetic polymorphism of drug metabolism. Indications for including a drug in a therapeutic drug-monitoring
program are:
.
There is an experimentally determined relationship between plasma drug concentration
and the pharmacological effect.
. There is a narrow therapeutic window.
. The toxicity or lack of effectiveness of the drug puts the patient at risk.
. There are potential patient compliance problems.
. The dose cannot be optimized by clinical observations alone.
Most drugs do not meet the criteria to be included in a TDM program. They are safely taken
without determining drug concentrations in plasma since the therapeutic effect can be
evaluated by other means. For example the coagulation time effectively measures the
Drug
dose
Plasma
concentration
Pharmacokinetic
variability
Concentration
at the receptor site
Pharmacologic
effect
Pharmacodynamic
variability
Figure 23.2 Effect of pharmacokinetics and pharmacodynamics on the dose–response
relationship
456
Introduction to Pharmaceutical Chemical Analysis
Table 23.1 Therapeutic range of common drugs subjected to therapeutic drug monitoring
Drug
Therapeutic range
Drug
Therapeutic range
Amitriptyline
Carbamazepine
Desipramine
Digoxine
Disopyramide
Ethosuximide
120–150 ng/ml
4–12 mg/ml
150–300 ng/ml
0.8–2.0 ng/ml
2–5 mg/ml
40–100 mg/ml
Nortriptyline
Phenobarbital
Phenytoin
Primidone
Theophylline
Valproic acid
50–150 ng/ml
10–40 mg/ml
10–20 mg/ml
5–12 mg/ml
10–20 mg/ml
50–100 mg/ml
efficacy of an anticoagulant drug and the blood pressure indicates the efficacy of a drug used
in the treatment of hypertension. In these situations it is best to adjust the dosage on the basis
of response.
The two major situations when TDM is advised is for drugs used prophylactically to
maintain the absence of a condition (depressive or manic episodes, seizures, cardiac
arrhythmias, organ rejection, asthma relapses) and the other to avoid serious toxicity for
drugs with a narrow therapeutic window. Examples are antiepileptic drugs, antidepressant
drugs, digoxin, phenytoin, theophylline, cyclosporine and HIV protease inhibitors and
aminoglycoside antibiotics. The therapeutic range of some drugs subjected to TDM is
shown in Table 23.1. It is essential that the methods used to determine drug concentrations
are not only selective for the drug and reliable, but that they are also well characterized and
fully validated.
23.1.3
Forensic and Toxicological Analysis
In toxicological analysis and in analysis of doping agents the analyte is unknown. Biological
samples are examined for the presence of compounds that are not normally present. In those
cases the analytical work is based on screening of unknowns, which is the analytical process
that takes place when no prior information about possible agents is available. The focus is on
drugs and their metabolites, narcotics and other substances that are toxicologically relevant.
Serious cases that are of criminal relevance may include:
.
.
Analysis of pharmaceuticals and addictive drugs that may impair human behavior;
Detection of poisons and evaluation of their relevance in determining causes
of death.
In the screening, sample preparation is directed towards extraction of compounds with
characteristic properties such as selective extraction of alkaline drugs, neutral drugs or
acidic drugs. The extracts are analyzed by temperature programming on a capillary column
in GC or by gradient elution in HPLC. Due to the serious legal consequences of forensic
cases, particular emphasis is placed on the quality and reliability of analytical results. The
work involves the application of at least two different analytical methods based on different
physical or chemical principles. Confirmatory analysis by gas chromatography/mass
spectrometry or liquid chromatography/mass spectrometry is normally required to interpret
analytical results for the court.
Analysis of Drugs in Biological Fluids
23.1.4
457
Doping Control Analysis
The World Anti-Doping Agency (WADA) was established in 1999 as an international
agency to promote, coordinate and monitor the fight against doping in sport. One of
WADA’s most significant achievements was the acceptance and implementation of the
World Anti-Doping Code (Code). The Code is the core document that provides the
framework for anti-doping policies, rules and regulations within sport organizations and
among public authorities. The Code works in conjunction with five International Standards
aimed at bringing harmonization among anti-doping organization in various areas. The
standards are:
.
Prohibited list;
International Standard for Testing;
. International Standard for Laboratories;
. International Standard for Therapeutic Use Exemptions;
. International Standard for the Protection of Privacy and Personal Information.
.
The prohibited list is the standard that defines substances and methods prohibited to athletes
at all times (both in competition and out of competition), substances prohibited in
competition and substances prohibited in particular sports. The prohibited list is updated
annually and falls into the categories shown in Box 23.1.
Box 23.1 Prohibited List in the World Anti-Doping Code
A. Substances and methods prohibited at all times:
1. Anabolic agents:
Anabolic androgenic steroids (AAS)
a. Exogenous anabolic androgenic steroids
b. Endogenous anabolic androgenic steroid
2. Other anabolic agents:
S2 Peptide hormones, growth factors and related substances
S3 Beta-2 agonists
S4 Hormone antagonists and modulators
S5 Diuretics and other masking agents
3. Prohibited methods:
M1 Enhancement of oxygen transport
M2 Chemical and physical manipulation
M3 Gene doping
B. Substances prohibited in competition:
S6 Stimulants
S7 Narcotics
S8 Cannabinoids
S9 Glucocorticosteroids
C. Substances prohibited in particular sports:
P1 Alcohol
P2 Beta-blockers
458
Introduction to Pharmaceutical Chemical Analysis
The purpose of the International Standard for Testing is to plan for effective incompetition and out of competition testing and to maintain the integrity and identity of
samples collected. The International Standard for Therapeutic Use Exemptions and the
International Standard for the Protection of Privacy and Personal Information ensures that
the process of granting an athlete therapeutic use exemptions is harmonized and that all
relevant parties adhere to the same set of privacy protections.
Only laboratories accredited by WADA takes part in the testing. The purpose of the
International Standard for Laboratories is to ensure that laboratories produce valid test
results. The standard further ensures that uniform and harmonized results are reported from
all accredited laboratories. In addition, the document specifies the criteria that must be
fulfilled by anti-doping laboratories to achieve and maintain their WADA accreditation.
There are currently 34 laboratories (2010) worldwide that are accredited by WADA to
conduct human doping control analysis.
To confirm the presence of a performance-enhancing drug or method according to the
prohibited list mass spectrometry coupled to either gas (GC) or liquid chromatography (LC)
is the method of choice. GC or LC coupled to MS or MS-MS are considered acceptable for
both initial screening procedures and for confirmation procedures for a specific analyte.
23.2 The Biological Matrix
Blood samples are collected to assess pharmacokinetics in drug development and to adjust
dosage in therapeutic drug monitoring. Urine samples, which are collected without
intervention, are the preferred matrix for screening of prohibited substances in doping
control analysis. Depending on circumstances different biological fluids and tissues are
examined in forensic and toxicological analysis.
The composition of blood is shown in Figure 23.3. Blood is composed of blood cells
suspended in a liquid called plasma. The blood cells are erythrocytes (red blood cells),
leucocytes (white blood cells) and thrombocytes (platelets). By volume the red blood cells
constitute about 46% of the blood volume for men and about 41% for women, plasma about
54–59% and the white blood cells about 0.7%. The portion of blood volume that is occupied
by red blood cells is called the hematocrit. Plasma is an aqueous solution containing about
8% of proteins and trace amounts of other substances such as dissolved nutrients and waste
Plasma
Plasma
54 – 59%
Leucocytes and
trombocytes
Erythrocytes
Blood cells
41 – 46%
Figure 23.3 The composition of blood
Analysis of Drugs in Biological Fluids
459
Table 23.2 Protein binding of common drugs
Drug
Digoxin
Captopril
Benzylpenicillin
Salicylic acid
Protein binding (%)
Drug
Protein binding (%)
20
30
50
80
Phenytoin
Imipramine
Ibuprofen
Warfarin
90
93
98
99
products, electrolytes, lipoprotein particles, immunoglobulins and blood clotting factors.
Blood pH is regulated to stay within a narrow range of 7.35–7.45.
Most drugs are bound to plasma proteins and exist in two forms, unbound or bound.
Common proteins that bind drugs are serum albumin and glycoprotein. The protein binding
is reversible and equilibrium exists between the bound and the unbound fraction:
Protein þ Drug $ Drug Protein complex
It is the free (unbound) fraction that exhibits pharmacologic effect. It is also the free fraction
that is metabolized and excreted. The bound fraction acts as a reservoir or depot from which
the drug is slowly released to maintain equilibrium. Acidic and neutral drugs are primarily
bound to albumin which is basic. Basic drugs will bind to acid a-1 glycoprotein. Table 23.2
shows the degree of protein binding of some common drugs. The protein binding is highly
dependant on the physicochemical properties of the drug. For example lipophilic drugs are
more protein bound that hydrophilic drugs. The protein binding may also vary between
persons, it may vary with age and it is affected by disease and by other drugs that bind to the
same protein. Sample preparation methods normally break the drug-protein bindings.
Bioanalytical methods therefore determines the total concentration of a drug in serum or
plasma and the result is unaffected by the protein binding.
When handling blood samples care must be taken to prevent hemolysis. Hemolysis is the
breakdown of red blood cells, which releases hemoglobin and other cell constituents into
plasma, which changes color from straw yellow to red. For this reason blood cells should be
removed as soon as possible and plasma or serum should be collected for analysis. Plasma or
serum is easier to handle than whole blood samples, and it is the concentration of a drug in
serum or plasma that is clinically relevant. When plasma is required, the blood sample is
collected in a clean tube containing an anticoagulant, and plasma is collected after
centrifugation (see Figure 23.3). When serum is required the blood sample is stored
until it is coagulated and serum is collected after centrifugation. Plasma can be prepared
rapidly after collection of blood samples and is often the preferred matrix for pharmacokinetic studies. Serum is isolated from coagulated blood and is the blood plasma with the
fibrinogens removed:
Serum ¼ Plasma Fibrinogens ðand other clotting factorsÞ
The test tubes mostly used for collection of plasma or serum samples are vacutainer tubes.
These are evacuated test tubes with a rubber cap. The vacutainer tubes are designed for
venipuncture and the blood is forced into the tubes by the vacuum. The tubes may contain
additional substances that preserve the blood for processing in the laboratory. Heparin is
frequently added as anticoagulant and is used for processing of plasma. Other
460
Introduction to Pharmaceutical Chemical Analysis
anticoagulants are potassium oxalate and EDTA (ethylenediaminetetraacetic acid). A clot
activator or a gel may be also added to vacutainer tubes for rapid serum separation.
Urine is an aqueous solution of byproducts of the body secreted by the kidneys. More than
95% of urine is water and the rest is a complex mixture of inorganic salts and organic
compounds. Some major constituents in the order of decreasing concentrations are urea
9.3 g/l, chloride 1.9 g/l, sodium 1.2 g/l, potassium 0.75 g/l and creatinine 0.67 g/l. In addition
to these substances urine contains a range of substances that vary with what is introduced
into the body (food). The composition of urine therefore varies between persons and during
the day. The pH of urine is normally close to 7 but can vary between 4 and 8. A diet of citrus
fruits and vegetables can increase urine pH and a diet of meat and cranberries can decrease
urine pH (more acidic). Drugs and chemicals can be taken to alter urinary pH. Intake of
sodium bicarbonate and potassium citrate makes urine pH more basic while intake of
ammonium chloride makes the urine more acidic. Urinary pH effects the excretion of a drug
into urine. For example, the excretion of an alkaline drug is greatly reduced in alkaline urine
because the solubility is reduced. Therefore the pH and the creatinine concentration are
normally checked during standard urine drug testing. High creatinine levels indicate a
normal sample, whereas low amounts of creatinine in the urine indicate a manipulated
(diluted) sample, either through the addition of water in the sample or by drinking excessive
amounts of water.
A variety of drugs are metabolized by conjugation with glucuronic acid or sulfate and are
excreted into urine as glucuronide and sulfate conjugates. These have a much higher water
solubility that the parent drugs. In urine drug-testing conjugates can be hydrolyzed to release
the parent drug and increase their concentration in the sample. For this purpose acid-base
hydrolysis or enzymes that cleave the glycoside bond of a glucuronide such as betaglucuronidase, are frequently used.
Normally, the concentrations of drugs and metabolites found in urine are higher than
those found in plasma or serum. Urine is therefore a preferred matrix for screening of
compounds that are not normally present such as doping agents, illicit drugs and poisons. In
addition urine sample are easily collected without intervention.
23.3 Bioanalytical Methods
Biological fluids are very complex and target analytes are often present at low concentrations. Chromatographic methods are the most important in bioanalytical work and these are
discussed in this chapter. Although bioanalysis has been performed with various instruments most of these cannot handle the biological fluid directly. Sample preparation is
usually necessary to isolate and extract the analytes. The various steps of a bioanalytical
method are shown in Figure 23.4 and include sampling, sample preparation, separation,
detection, identification, calibration and quantification.
23.3.1
Sampling
Depending on circumstances blood samples are collected at fixed time points and plasma
or serum are separated as soon as possible. Whole blood may be the sample of choice for
Analysis of Drugs in Biological Fluids
Sampling
Sample
preparation
Separation
461
Detection
Identification/Quantification
Calibration
Figure 23.4 The steps of a bioanalytical method
detection of an analyte in forensic toxicology when processing of serum or plasma is not
possible because of hemolysis. The volumes collected should be kept as small as
possible and range from a few microliters from laboratory animals up to milliliter
volumes (5 ml) from humans. When immediate analysis is impossible, serum and plasma
samples are normally stored in a freezer at 21 C prior to analysis. The stability of the
analyte must be documented during storage and in the following freeze–thaw process.
There are only minor differences between serum and plasma and methods developed
for plasma are usually applicable for serum. Hemolyzed whole blood samples are
more complex and require extensive sample preparation. The collection and handling
of blood and other biological fluids requires great care because of possible presence
of transferable disease constituents. Gloves and protective eyeglasses must be worn at
all times.
Urine is collected in clean sample bottles. In doping control 90 ml of urine are required
and urine samples can be stored in a refrigerator for a short period of time. Urine from
healthy persons is sterile and is relatively free from protein. When properly handled and
stored, the risk of bacterial degradation is greatly reduced. The time between sampling and
analysis should be kept as short as possible to guarantee stability. If necessary, preservatives
may be added to stabilize the analyte. Urine may contain compounds that can be irritating to
eyes and skin. Gloves and protective eyeglasses are recommended when handling urine
samples in the laboratory.
23.3.2
Sample Preparation
Sample preparation is crucial for successful bioanalysis and constitutes about 80% of the
total analysis time. The major goals of sample preparation are:
.
Removal of unwanted matrix components;
Concentration of an analyte to improve detection limits;
. Exchange of solvent from the aqueous biological fluid to a solvent more suitable for
injection into the chromatographic system.
.
For analysis of drugs in biological fluids and tissues, proteins must usually be removed from
the sample matrix before a sample solution is injection into a chromatographic system. If
serum or plasma were injected into a reversed phase HPLC system, the consequences may
be precipitation of the proteins by the mobile phase, clogging of frits and lines resulting in
462
Introduction to Pharmaceutical Chemical Analysis
rapid increase of the backpressure of the system and detoriation of the chromatographic
column. In addition to proteins it may also be desirable to remove other matrix components
that may interfere with the analysis such as small molecules and metabolic byproducts.
When analyzing basic (acidic) analytes it may be desirable to remove neutral and acidic
(basic) compounds to increase the specificity of the method. It may also be necessary to
remove selected compounds when the resolving power of the system is insufficient to
separate the compounds.
Biological fluids may also contain analyte concentrations that are below the limit of
quantification. In these situations it is necessary to concentrate the analyte to improve
detection limits in quantitative determinations. The analyte is then extracted and reconstituted in a smaller amount of solvent prior to injection into the chromatographic system. If
the analyte is quantitatively extracted from 1.0 ml of biological fluid and reconstituted in
0.1 ml of solvent the detection limit has increased with a factor of 10.
The solution injected into a chromatographic system should be compatible with the
mobile phase and exchange of solvent from the aqueous biological fluid to a solvent more
suitable for injection is a common goal in sample preparation. A method which combines
solvent exchange and analyte concentration is to evaporate a given volume of an extract to
dryness, then reconstitute in a smaller volume of solvent for example the mobile phase in
HPLC or a volatile organic solvent for GC analysis.
23.3.3
Protein Precipitation
The most common way to remove proteins from serum and plasma is by precipitation. The
fundamentals of protein precipitation are discussed in Section 18.4. Common protein
precipitation procedures are based on denaturation by acids or by water-soluble organic
solvents. An excess of precipitating reagent is added to plasma or serum, the proteins are
precipitated and the supernatant is collected after centrifugation. When the concentration of
analyte in the supernatant is above the quantification limit and separation and detection is
performed without interferences, no further concentration and cleanup steps are necessary
and the solution can be injected into a HPLC system. The characteristics of common protein
precipitation reagents are shown in Table 18.1.
Water-miscible organic solvents such as acetonitrile or methanol are used most frequently for precipitating the proteins, because of the excellent solubility properties of the
solvent–water mixture and because of the compatibility of the solvent mixture with
common HPLC mobile phases. The mild conditions used minimize the possibility for
decomposition of labile drug substances and normally near complete recoveries ( > 90%)
are obtained because most drugs are very soluble in the organic solvent-water mixture. As
shown in Table 18.1 a solvent–plasma ratio of 2 : 1 has to be used to precipitate more than
98% of the proteins. The dilution of the sample solution is considered a disadvantage
because it reduces detection limits.
As shown in Table 18.1 a 10% (w/v) solution of trichloroacetic acid (TCA) is the most
powerful deprotonization reagent and 0.2 ml of the reagent precipitates more than 99% of
the proteins from 1 ml of plasma. The supernatant is highly acidic. Acidic drugs are poorly
soluble in the acidic solvents and recoveries are low. Drugs can also coprecipitate with the
proteins which reduces recoveries. Alkaline drugs are normally recovered with high
recoveries because they are highly soluble in acidic aqueous media.
Analysis of Drugs in Biological Fluids
463
As only proteins are removed by protein precipitation the low molecular components of
plasma or serum are still present in the supernatant. When analyte concentrations are
determined by HPLC efficient separation systems and specific detection are necessary to
eliminate possible interferences from matrix components. Protein precipitation is a
preferred sample preparation method in drug discovery, because of the simplicity of the
method and because the method is easily automated in laboratory workstations.
Extraction methods such as liquid–liquid extraction (LLE) or solid–phase extraction
(SPE) are preferred when sample cleanup, analyte enrichment or solvent exchange
is required. These methods are described in Chapter 18 and are applicable to all
biological fluids.
23.3.4
Liquid–Liquid Extraction
Liquid–liquid extraction is the traditional sample preparation technique that depends on the
partitioning of the analyte between two immiscible solvents. The fundamentals of LLE are
discussed in Section 18.5. Analytes distribute between the aqueous phase and the organic
solvent and hydrophobic drugs partition into the organic phase when the drugs are unionized.
Isolation of the organic phase, evaporation of the solvent followed by reconstitution in a
solvent compatible with the chromatographic technique yields a sample ready for injection.
Many methods have been developed for the extraction of drugs from biological fluids. For
extraction of basic drugs, the sample is made alkaline to neutralize the drug, and for extraction
of acidic drugs the sample is acidified to neutralize the drug. When selecting a suitable solvent
volatility, polarity, selectivity, solubility and density must be considered. Solvents that are
harmful to the health and the environment should not be used. Evaporation of solvent is
facilitated by a solvent with high volatility. Solvent polarity is important for the solubility of
the drug and for the partitioning of the drug into the organic phase. Low solubility of the
matrix components reduces possible interferences problems. Solvents with densities lower
than water stay on top of the two-phase extraction system and are easily collected. Common
solvents used for drug analysis and their boiling points are shown in Table 18.2.
Normally a single extraction is performed and the volume of solvent is adjusted to achieve
a satisfactory recovery. With proper selection of organic solvent and adjustment of pH clean
extracts can be obtained. Inorganic salts, proteins and hydrophilic matrix components
remain behind in the aqueous phase and hydrophobic neutral compounds partition into the
organic phase. When necessary, hydrophobic neutral compounds can be removed by back
extraction. Basic drugs can be back extracted into an acidic aqueous solution and acidic
drugs can be back extracted into an aqueous alkaline solution. Neutral hydrophobic
substances remain in the organic phase.
The disadvantages of LLE include safety when handling organic solvents. LLE methods
are labor-insensitive and difficult to automate. However, despite of several drawbacks LLE
is widely used because of the efficient cleanup and the high enrichment obtained and
because LLE methods are easily standardized.
23.3.5
Solid-Phase Extraction
Solid-phase extraction (SPE) is an alternative technique that eliminates most of the
drawbacks associated with LLE. The fundamentals of SPE are discussed in Section
464
Introduction to Pharmaceutical Chemical Analysis
18.6. SPE is based on partitioning of the analyte between the aqueous biological fluid and a
solid phase. Intermolecular forces between the analyte and the two phases determine analyte
retention and elution. The solid phase are sorbent particles contained in a column or disc.
Sorbent particles packed into syringe barrels are used most frequently as SPE columns in
combination with vacuum manifolds. The amount of sorbent particles determines extraction
capacity and the volume of the SPE column depends on the sample volume: 1-ml columns
packed with 30–100 mg of sorbent are normally used for extraction of 1 ml of plasma or
serum. The columns are normally discarded after extraction. Reversed phase, ion exchange
and mixed mode sorbents are used for extraction of biological fluids. Conditioning is by
methanol or acetonitrile, followed by water or an aqueous buffer. Biological fluids are
normally diluted with an aqueous buffer prior to loading. Dilution decreases sample
viscosity and the solution is sucked more rapidly through the sorbent bed. The pH can
be adjusted to increase analyte retention. Matrix components not absorbed by chemical
attraction remain in the aqueous phase and go to waste. In reversed phase extractions
hydrophilic matrix components and most of the proteins go to waste. The wash solvent
should remove contaminating matrix components, while the analytes are retained in the
sorbent bed. Analyte elution is normally with methanol or acetonitrile often modified with
an acid or a base. The elution solvent must have good solubility properties for the
analyte and should effectively break the chemical bindings between the analyte and the
sorbent. Solvent change is often performed prior to injection into a chromatographic system
because methanol and acetonitrile are too strong eluents to be injected directly into reversed
phase HPLC.
The advantages of SPE include a great reduction in the consumption of solvents, shorter
sample preparation times, and simultaneous preparation of several samples and ease of
automation. Many sorbents and formats are available to meet most needs. One drawback is
the cost per sample that is higher than with LLE.
23.3.6
Separation
HPLC is the standard separation method in drug discovery and in therapeutic drug
monitoring. UHPLC methods are used increasingly because of high speed, reduced solvent
consumption and lower detection limits. Reversed phase separations on C18 phases are most
common. GC with temperature programming on nonpolar or medium-polar capillary
columns is standard for the screening of prohibited substances in doping control analysis.
Sample preparation procedures include derivatization to increase the volatility of target
analytes, to improve the chromatographic separation and to lower detection limits. In
forensic toxicology both HPLC and GC are used as screening methods. GC methods are
more labor-intensive and time-consuming than HPLC methods. For the screening of
selected compounds HPLC methods based on gradient elution on reversed phase columns
offer an alternative to GC methods.
23.3.7
Detection
The mass spectrometer is the detector of choice for most applications. The mass spectrometer provides mass spectra which, combined with retention data from the chromatographic
separation, are used to identify unknown substances. Recording the intensities of masses
Analysis of Drugs in Biological Fluids
465
that are characteristic for target analytes is the basis of quantification. The disadvantages of
mass spectrometers are the high costs of the instruments and the high operating costs. For
certain applications less specific detectors can be used such as UV, fluorescence or
electrochemical detectors in HPLC and flame ionization, nitrogen-specific or electron
capture detectors in GC.
23.3.8
Calibration and Quantification
Calibration can be with external or internal standards. Internal standards are normally
preferred because higher reliability of data is obtained. External standard calibration can be
used when few sample preparation steps are necessary such as in protein precipitation from
serum or plasma. A reproducible analyte recovery, control of sample volume and control of
injection volume is required. Test tubes must be closed with a cap to eliminate possible
evaporation of solvent prior to injection into a HPLC system.
An internal standard calibration is used when the analyte is extracted. An appropriate
internal standard controls for extraction variability and for variability in injection volumes.
When MS is used as detector the best internal standard is an isotopically labeled version of
the analyte. Isotopically labeled substances, however, are expensive. Typically, deuterated
analogs are used, containing 3–5 deuterium atoms. Other stable isotopes can also be used,
such as 13 C and 18 O . Therefore compound analogs that have similar extraction recoveries
and similar chromatographic and detection characteristics as the analyte are mostly used as
internal standards.
Quantification is based on standard curves set up after analysis of standards solutions.
Standard solutions are prepared from drug-free biological fluid. Standard solutions for
plasma analysis are prepared in drug-free plasma and standard solutions for urine are
prepared in drug-free urine. Standard solutions containing known concentrations of the
analyte are prepared from stock solutions. The concentration range should be the same as the
expected concentration range of the target analyte. Normally stock solutions of known
concentrations are prepared in an aqueous solution or in water miscible organic solvents
(methanol or acetonitrile). A small volume of the stock solution is then mixed with a fixed
volume of the drug-free biological fluid to prepare standard solutions of known concentrations. Prior to analysis standard solutions are stored at the same conditions as the
sample solutions.
Internal standards should be added to sample solutions and to standard solutions at the
beginning of sample preparation, typically before addition of a protein-precipitating reagent
or before extraction. Normally the same volume of the internal standard stock solution is
added to keep the concentration of the internal standard constant. Standard solutions and
sample solutions are treated equally in the sample preparation process. Then the same
fraction of the analyte and the internal standard is eventually lost, and the ratio of analyte to
internal standard response remains constant.
Prior to quantitative determinations an extract from a drug-free biological fluid should
be analyzed to demonstrate that the sample preparation process has successfully removed
possible interferences. The detector response ratio of the analyte and internal standard is
calculated after analysis of standard solutions and sample solutions. The response ratio
can be peak height ratio or peak area ratio. When the same amount of internal standard is
added to standard solutions and to sample solutions the calibration curve is a plot of
466
Introduction to Pharmaceutical Chemical Analysis
response ratio against concentration. Unknown concentrations are determined from the
standard curve.
23.4 Examples
The following examples discuss various sample preparation methods for amitriptyline in
serum, the quantitative determination of amitriptyline in serum by LC-MS, the quantitative
determination of valproic acid in serum by GC-MS and the identification of morphine in
urine by GC-MS and LC-MS-MS.
23.4.1
Sample Preparation
Amitriptyline (Figure 23.5) is one of the most widely used tricyclic antidepressant drugs and
is frequently included in TDM programs. In the following, sample preparation procedures
for amitriptyline in serum based on LLE, SPE and PP are discussed.
23.4.1.1 Sample Preparation Procedure by LLE
1 ml of serum is transferred to a centrifuge tube and added to 50 m of an internal standard
solution and 500 ml of 0.75 M sodium bicarbonate/carbonate buffer, pH 10, followed by 5 ml
of methyl tert-butyl ether – n-hexane (75 : 25, v/v). The mixture is shaken for 30 s on a vortex
mixer and then centrifuged at 4000 rotations per minute (rpm) for 10 min to separate the
aqueous and the organic phase. The organic phase is transferred to another test tube and
evaporated to dryness under a stream of nitrogen at 40 C. The residue is reconstituted in
100 ml of mobile phase prior to injection into the HPLC system.
23.4.1.2 Comments to the Procedure
The internal standard is an analog of amitriptyline and is added first to lower measurement
uncertainty. Amitriptyline is a hydrophobic base with a pKa value of 9.4 and carbonate
buffer pH 10 is added to neutralize the drug enabling the uncharged drug to be extracted into
the organic phase. The addition on n-hexane to methyl tert-butyl ether increases the
hydrophobicity of the solvent mixture. Only neutral hydrophobic substances (including
Chemical Formula: C20H23N
Exact Mass: 277.18
Molecular Weight: 277.40
pKa = 9.4
log P octanol (pH 7.4) = 3.0
N
Figure 23.5 Chemical structure and properties of amitriptyline
Analysis of Drugs in Biological Fluids
467
the analyte and the internal standard) are extracted into the organic phase. The polar
constituents of serum and acidic compounds remain in the aqueous phase. It was shown in
method development that the addition of hexane increased the specificity as the extraction of
more polar constituents of plasma was reduced and that near complete recovery of
amitriptyline was obtained with 5 ml of the solvent mixture. Mixing is by the vortex
mixer and 30 s of mixing was sufficient. However, because of the powerful mixing highspeed centrifugation for 10 min is necessary to separate the phases. The solvent mixture has
a low boiling point and is easily evaporated at 40 C. This procedure includes removal of
unwanted matrix components, ten times enrichment as the analyte is extracted from a 1 ml
sample and reconstituted in 100 ml of mobile phase and solvent exchange.
23.4.1.3 Sample Preparation Procedure by LLE and Back Extraction
1 ml of serum is added to 50 ml of an internal standard solution and 500 ml of 0.75 M
sodium bicarbonate/carbonate buffer pH 10 followed by 8 ml of heptane-isopentyl alcohol
(98 : 2, v/v). The mixture is shaken for 5 min, centrifuged and set in a freezer at 21 C until
the aqueous phase is frozen. The organic layer is removed and added to 150 ml of 0.025 M
phosphate buffer pH 2.5. The mixture is shaken for 5 min and the aqueous phase is collected
and an aliquot is injected into the HPLC system.
23.4.1.4 Comments to the Procedure
The solvent mixture used to extract amitriptyline from serum has a higher boiling point than
the solvent mixture used to extract amytriptyline in the previous procedure. In this
procedure solvent evaporation is eliminated and a higher boiling point is recommended
as it reduces evaporation of solvent in the sample preparation process. Some emulsion
formation may occur in the extraction and complete removal of the organic phase may be
difficult. Freezing which transforms the aqueous phase into ice reduces this problem.
Amitriptyline is protonated at pH 2.5 and is collected in phosphate buffer after back
extraction. Serum has a high buffer capacity and 500 ml of 0.75 M sodium bicarbonate/
carbonate buffer was necessary to maintain a constant pH of 10 during the extraction of
amitryptyline from serum. In the back extraction from the heptane–isoamyl alcohol phase
there is no need for a high buffer capacity and a 0.025 M phosphate buffer is sufficient to
maintain a constant pH of 2.5 in aqueous phase. Only hydrophobic basic compounds are
back extracted to the acidic aqueous phase. Neutral hydrophobic compounds remain in the
organic phase. This procedure includes removal of unwanted matrix components, analyte
enrichment and solvent exchange.
23.4.1.5 Sample Preparation Procedure by SPE
The SPE column is a 1-ml column packed with 100 mg C18 sorbent particles. Vacuummediated elution is used and the flow rate is adjusted to 1 ml/min. The columns are
conditions with 2 1 ml of methanol followed by 1 ml of water. After conditioning, 0.5 ml
of plasma is added to 100 ml of the internal standard solution and 0.5 ml of water is then
sucked through the column. The column is washed with 2 1 ml of water and with 2 1 ml
of 25% metanol in water. The analyte is eluted with 0.5 ml of methanol added 1% (v/v)
perchloric acid. An aliquote of the eluate is injected into the HPLC system after 1 : 1 dilution
with water.
468
Introduction to Pharmaceutical Chemical Analysis
23.4.1.6 Comments to the Procedure
In its dry form, the C18 bonded phase is randomly oriented on the surface of the particles
forming clusters or aggregates among bonded groups that are close to each other. In such a
configuration, the hydrophobic surface that is exposed to an analyte is very small. Therefore,
the nonpolar SPE sorbent is first conditioned with methanol. With an environment of
methanol surrounding the C18 groups the sorbent becomes more open and available for
interaction with the analyte. Excess methanol held between sorbent particles and in the void
volume of the SPE column is undesirable and is washed out with water. Amitriptyline,
which is a hydrophobic base, is retained by hydrophobic interaction to the C18 groups. It can
also bind to residual silanol groups by secondary interaction. At a neutral pH residual silanol
groups are negatively charged and positively charged compounds can be retained by ionic
interaction. After the sample solution is drawn through the column by vacuum the column is
first washed twice with water to remove traces of proteins and other polar components in
serum. Washing with 25% methanol in water removes compounds that are more polar and
less retained than amitryptiline. Amitryptyline is eluted with methanol added 1% perchloric
acid that effectively breaks primary and secondary interactions. Perchloric acid neutralizes
residual silanol groups and methanol releases drug molecules retained by hydrophobic
interaction. This procedure includes removal of unwanted matrix components and solvent
exchange. Evaporation and reconstitution in a smaller volume is necessary when enrichment is necessary.
23.4.1.7 Sample Preparation Procedure by Protein Precipitation
1 ml of serum is transferred to a centrifuge tube and 2.0 ml of acetonitrile are added. The
mixture is shaken for 5 min and centrifuged at 10 000 rpm for 10 min. 0.5 ml of the
supernatant is transferred to a test tube and an aliquot is injected into the HPLC system.
23.4.1.8 Comments to the Procedure
This procedure removes most of the proteins in the serum sample. The low molecular
weight constituents are all present in the supernatant. There is no solvent exchange and the
sample solution has been diluted. No internal standard is added and quantification is based
on external standards. A reproducible analyte recovery, control of sample volume and
control of injection volume is required. Centrifuge tubes and test tubes must be closed to
eliminate evaporation. The major advantage of protein precipitation as compared to
procedures based on LLE or SPE is the simplicity of the method. Efficient separation and
specific detection are necessary to eliminate interferences from matrix components. When
solvent exchange or enrichment is required the supernatant is evaporated and reconstituted in a smaller volume of solvent. In this case an internal standard is usually added prior
to protein precipitation.
23.4.2
Quantitative Determination
23.4.2.1 Quantitative Determination of Amitriptyline in Serum by LC-MS
Principle. The serum sample is added to an internal standard and extracted by the reversed
phase SPE procedure described previously. The compounds are separated on a reversed
phase column and detected by MS in the selected ion monitoring (SIM) mode. Quantification is based on measuring peak area ratios of amitriptyline and the internal standard.
Analysis of Drugs in Biological Fluids
469
Procedure. The analytes are separated on a UHPLC column (50 2.1 mm) column packed
with 1.7-mm C18 particles. The column temperature is 40 C and the mobile phase is
acetonitrile with 0.01 M ammonium acetate (85 : 15, v/v; pH 3.5) delivered at a flow rate of
0.3 ml/min. 10 ml of the extract is injected into the UHPLC column. Compounds eluted from
the UHPLC are directed into the electrospray ion source operating in the positive mode. SIM
mode is used for quantification by the protonated molecular ions (MþH)þ at m/z 278 for
amitriptyline and at m/z 315 for the internal standard. The retention times are 55 and 70 s,
respectively. Quantification is based on a calibration curve generated from peak area ratios of
standards of amitriptyline (50–200 ng/ml) and the internal standard (100 ng/ml) in plasma.
23.4.2.2 Comments to the Procedure
Reversed phase LC separations of basic drugs are usually performed with an acidic mobile
phase. The reason for this is that silica based columns usually degrade at high pH. Mobile
phases must be volatile in order to be compatible with the mass spectrometer. Therefore a
volatile ammonium acetate buffer is used to adjust the pH of the mobile phase to 3.5. Fast
separation is achieved by UHPLC and total analysis time is 3 min. Amitriptyline and the
internal standard have molecular masses (M) of 277 and 314 and the (MþH)þ ions are
detected in the mass spectrometer in the SIM mode.
Figure 23.6 show chromatograms of amitriptyline detected at m/z 278 (MþH)þ and of the
internal standard detected at m/z 315 (MþH)þ . Selective ion monitoring at m/z 278 and m/z
0
(a)
m /z 278 (M+H)+
(b)
m/z 315 (M+H)+
1
2
3
min
Figure 23.6 Chromatograms of amitriptyline (a) and the internal standard (b) detected at
m/z 278 and m/z 315
470
Introduction to Pharmaceutical Chemical Analysis
Table 23.3 Peak areas and peak area ratios of standards of amitriptyline (50–200 ng/ml) and the
internal standard (100 ng/ml) in plasma
Concentration of
amitriptyline (ng/ml)
Peak area of
amitriptyline
Peak area of
internal standard
Peak area
ratio
126 398
248 985
378 236
498 168
257 363
255 487
258 134
256 874
0.491127
0.974551
1.46527
1.939348
50
100
150
200
315, respectively, showed no other peaks and demonstrated the selectivity of the UHPLCMS procedure.
Standard solutions containing 50, 100, 150 and 200 ng/ml of amitriptyline and 100 ng/ml
of the internal standard were prepared in drug-free plasma. These were analyzed according
to the procedure above and the peak areas of amitriptyline and the internal standard are
shown in Table 23.3.
The same amount of internal standard was added to standard solutions and to sample
solutions, and the calibration curve is a plot of peak area ratios against concentration as
shown in Figure 23.7. The regression line y ¼ 0.0097x þ 0.0035 [where y is the peak area
ratio and x is the concentration of amitriptyline (in ng/ml)] is used for the calculation of
unknown concentrations. In the chromatogram of an unknown sample the peak areas for
amitriptyline and the internal standard were 354 048 and 255 380 respectively.
The peak area ratio was 354 048/255 380 ¼ 1.3863. The unknown concentration, x, was
calculated from the regression line:
1:3863 ¼ 0:0097x þ 0:0035
x ¼ ð1:3863 0:0035Þ=0:0097 ¼ 142:6
2.5
Peak area ratio
2
1.5
1
y = 0.0097 + 0.0035
0.5
0
0
50
100
150
200
ng/ml
Figure 23.7 Calibration cure based on internal standard
250
Analysis of Drugs in Biological Fluids
471
The concentration of amitriptyline in the sample was 143 ng/ml.
23.4.2.3 Determination of Valproic Acid in Serum by GC-MS
Principle. The internal standard is an isotopically labeled version of the analyte where
16
O have been replaced with 18 O . Both compounds are extracted from serum samples by
liquid–liquid extraction and the extract is analyzed by GC-MS by SIM after derivatization.
Quantification is based on measuring the peak area ratio of nonlabeled to labeled valproic
acid. The structures of valproic acid and the labeled internal standard are shown in
Figure 23.8.
Procedure. 0.2 ml of plasma is added 0.1 ml of a solution containing 100 mg of the
internal standard. The mixture is added 0.1 ml of 1 M HCl and 0.5 ml of hexane. After
vortex mixing for 1 min the phases are separated by centrifugation. 0.2 ml of the hexane
phase is collected and 0.05 ml of derivatizing reagent comprising 1% trimethylchlorosilane-pyridine (2 : 1, v/v) dissolved in N,O-bis(trimethysilyl)-trifluoroacetamide
(BSTFA) is added. 1 ml of the solution is injected into GC-MS.
The injector is a split injector operated at a split ratio of 1 : 10 and the injector temperature
was 260 C. The column is a 15-m capillary column with an inner diameter of 0.25 mm. The
stationary phase is phenyl-methyl-siloxane. Helium is used as carrier gas at a flow rate
of 1.5 ml/min. The column temperature is programmed from 100 to 250 C at 40 C/min.
The compounds are detected by SIM at mass m/z 201 for the derivative of valproic acid and
at m/z 205 for the derivative of the labeled internal standard.
23.4.2.4 Comments to the Procedure
Valproic acid has a pKa value of 4.8 and is present as an anion in plasma. Addition of
hydrochloric acid neutralizes the analyte and facilitates the extraction into hexane. Valproic
acid is converted to the trimethylsilyl (TMS) derivative prior to GC analysis as shown in
Figure 23.9.
18
16
O
O
16
18
OH
OH
Valproic acid
Chemical formula: C8H16O2
Exact mass: 144.12
Molecular weight: 144.21
Internal standard
Chemical formula: C8H16O2
Exact mass: 148.12
Molecular weight: 148.21
Figure 23.8 Structure of valproic acid and the isotopically labeled internal standard
472
Introduction to Pharmaceutical Chemical Analysis
O
O
CH3
O
OH
Si
Silylation
CH3
H 3C
Figure 23.9 Derivatization of valproic acid
The derivatization increases analyte volatility and improves the chromatographic properties. The isotopically labeled internal standard has the same extraction recovery, chromatographic retention and ionization response as the analyte in the mass spectrometer. The
quantitative determination is based on selected ion monitoring of valproic acid at m/z 201
and at m/z 205 for the internal standard. The standard curve is a plot of peak area ratio of
valproic acid to the internal standard as described previously.
23.4.3
Identification
Traditionally, urine has been the sample of choice for screening and identification of
unknown drugs or poisons. Urine sampling is physically non-invasive, and large volumes
are available. Enzymatic hydrolysis of urine samples using b-glucuronidase/arylsulfatase is
frequently used followed by extraction to isolate alkaline substances, acidic substances and
neutral substances.
23.4.3.1
Sample Preparation Procedure for Unknown Screening by Mixed Mode Cation
Exchange
1 ml of urine samples is mixed with 2 ml of 0.1 M acetate buffer, pH 4.8. The mixture is
added to 25 ml b-glucuronidase/arylsulfatase and enzymatically deconjugated for 2 h at
50 C and then cooled to room temperature. The SPE column is a 10-ml column packed
with 130 mg mixed mode sorbent with C8 and sulfonic acid groups. Vacuum-mediated
elution is used and the flow rate is adjusted to 2 ml/min. The columns are conditioned with
3 ml of methanol followed by 3 ml of 0.25 M acetate buffer, pH 4.8. The hydrolyzed urine
samples are drawn through the columns. The columns are washed with 1 ml of water and
with 1 ml of 0.01 M hydrochloric acid. Before elution, the columns are dried by vacuum.
Elution at 1 ml/min involves 1 ml of methanol and 1 ml of methanol/ammonia (98 : 2 v/v).
The two fractions are collected separately and are evaporated to dryness in a stream of
nitrogen at 40 C.
23.4.3.2 Comments to the Procedure
In mixed mode extraction, the sorbent contains both hydrophobic and ionizable groups,
which makes it possible to retain urine constituents by hydrophobic interactions and by ion
exchange interactions. The column is usually conditioned with methanol and water. The pH
Analysis of Drugs in Biological Fluids
473
is adjusted to 4.8 with acetate buffer in the sample solution and in the conditioning step. A
weakly acidic pH in the sample solution promotes the enzymatic hydrolysis and keeps basic
compounds ionized and reduces the ionization of acidic compounds. When the sample is
loaded, basic compounds are retained by ionic interactions to the negatively charged
benzene sulfonic acid groups and neutral and acidic compounds are retained by hydrophobic interactions to the C8 groups. Washing with water removes the enzymes and unwanted
salts. Washing with 0.01 M HCl regenerates the capacity of the cation exchanger, which may
have lost capacity when the sample was loaded. The acidic pH maintains the ionization of
basic substances. Traces of aqueous wash solutions are removed by vacuum prior to elution
with an organic solvent. A general fractionation is performed in which acidic/neutral
components are eluted in one fraction and basic compounds are eluted in a second fraction.
Methanol elutes the neutral and acidic compounds that were retained by hydrophobic
interaction. Basic compounds, which are retained by ionic and hydrophobic interactions, are
eluted with methanol–ammonia. Ammonia is added to neutralize the cationic retention
mechanism and methanol-ammonia breaks all the interaction between the sorbent and the
basic analytes. Evaporation to dryness has the advantage that, upon reconstitution, one can
choose an organic solvent that is optimal for the subsequent chromatographic separation.
Capillary GC coupled to MS has traditionally been a reference method for identification of unknown substances in biological fluids. This example focuses on the identification of drugs of abuse in urine. Sample preparation was with mixed mode extraction
as described previously. The fraction containing alkaline drugs was subjected to
chromatographic analysis.
23.4.3.3 GC-MS Procedure for Unknown Screening
Principle. Compounds are converted to trimethylsilyl (TMS) ethers by derivatization and
separated on a capillary column by temperature programming. Detection is with mass
spectrometry operated in the full scan mode. Identification is based on retention data and the
full scan mass spectra.
Procedure. The extract is evaporated to dryness and the dry residue is derivatized with a
mixture of N,O-bis(trimethysilyl)-trifluoroacetamide–trimethylchlorosilane (BSTFATMCS; 99 : 1, v/v) at 70 C for 30 min and 2 ml is injected into the chromatograph in the
splitless mode. A 12-m capillary column with 0.2-mm inner diameter coated with phenylmethylsiloxane is used. Helium is used as carrier gas at a flow rate of 1.5 ml/min. The injector
temperature is 240 C. The column temperature is held at 90 C for 1 min followed by a 30 C
ramp to 200 C and a 5 C ramp to 250 C. The mass spectrometer is operated in the full scan
mode recording masses in the m/z range 40 to 500. A component elutes with a retention time
of 12.5 min, with the same retention time as morphine. The full scan mass spectrum is shown
in Figure 23.10 and a computer search shows the mass spectrum is identical to the spectrum of
bis-trimethyl silyl derivative of morphine. The identification is confirmed by analyzing a
standard solution of morphine according to the procedure described above.
23.4.3.4 Comments to the Procedure
The TMS derivatization is carried out to improve the chromatographic properties of polar
analytes. The derivatization converts the polar –OH groups of morphine into nonpolar –
OTMS groups as shown in Figure 23.11.
474
Introduction to Pharmaceutical Chemical Analysis
Figure 23.10 Mass spectrum of morphine di-TMS derivative
Compounds are separated on the capillary column by temperature programming and
mass spectra are collected for each compound as they exit the column. In the total ion current
(TIC) mode, the sum of all fragments ions in a GC peak is plotted against retention time. The
mass spectrum of each molecule is stored in the systems computer and the mass spectrum
CH3
H 3C
Si
O
HO
H3 C
Silylation
O
O
H
H
N
N
H3 C
O
HO
Si
H 3C
CH3
Chemical formula: C17H19NO3
Exact mass: 285.14
Molecular Weight: 285.34
Chemical formula: C23H35NO3Si2
Exact mass: 429.22
Molecular Weight: 429.70
Figure 23.11 TMS derivatization of morphine
Analysis of Drugs in Biological Fluids
475
corresponding to a given GC peak can be read out. Manufacturers of mass spectrometers
provide computer libraries of mass spectra of thousands of compounds, and special
computer searches can be made to match an unknown spectrum.
The mass spectrum of the peak eluted at the retention time of morphine is shown in
Figure 23.10 and the computer search showed that the spectrum matched the spectrum of the
di-TMS derivative of morphine. The fragment ion at m/z 429 corresponds to M þ ion of the
TMS derivative of morphine. The final identification was based on injection of the TMS
derivative of morphine standard. A complete match in the GC retention time and mass
spectra of the standard confirmed the identification.
Although GC-MS is a well established sensitive and specific technique, GC-MS analysis
usually requires time-consuming sample preparation steps, which includes derivatization of
the analytes. Derivatization is unnecessary in LC-MS and the chromatographic run times are
normally shorter. LC-MS has therefore replaced GC-MS for several applications. The
example shown below is an application of LC-MS-MS for the identification of morphine in a
mixed mode urine extract.
23.4.3.5 LC-MS-MS Procedure for Unknown Screening
Principle. Compounds are separated by gradient elution on a reversed phase UHPLC
column. Detection is with MS-MS operated in the selected reaction monitoring (SIR) mode.
Identification is based on retention data and MS-MS spectra.
Procedure. The alkaline extract from the mixed-mode SPE is evaporated at 40 C
using a flow of nitrogen gas and the residue is reconstituted in 200 ml water–acetonitrile
(95 : 5, v/v). Separation of the sample is performed on a UHPLC system at 35 C using a
C18 column (100 2.1 mm i.d., 1.7 mm particle size). Mobile phase A is 0.1% formic acid
and mobile phase B is acetonitrile–0.1% formic acid (9 : 1, v/v). A stepwise gradient at a
flow of 0.4 ml/min starts at 0% B. From 1 to 4 min the % of B increases linearly to 40% and
further increases linearly to 100% B in 10 min with a final hold for 2 min. The total run to
run time (including re-equilibration prior to injection of the next sample) is 13 min. The
injection volume is 20 ml.
The effluent from the LC is passed directly into the ion source of the MS-MS instrument
operating in the positive electrospray mode. In the selected reaction monitoring mode
(SRM), the precursor ion for detection of morphine is the protonated molecular ion,
(MþH)þ at m/z 286. Subsequent MS-MS fragmentation of the precursor ion is carried out
and the product mass at 201, which is characteristic for morphine, is measured. The peak
eluting at a retention time similar to morphine is shown to match the retention time and the
SRM chromatogram of an authentic morphine standard.
23.4.3.6 Comments to the Procedure
The gradient staring at 0% B and ending with 100% B is optimized to separate common
drugs of abuse. In the start no organic modifier is present in the mobile phase and
hydrophobic compounds are completely focused on the column giving narrow peaks in
the chromatographic separation. The run time of the UHPLC system is short and new
samples can be injected every 13 min. The precursor ion selected in the first quadrupole of
476
Introduction to Pharmaceutical Chemical Analysis
the MS-MS instrument is the protonated molecular ion (MþH)þ . Morphine has a molecular
mass of 285 and the precursor ion is at m/z 286. The precursor ion is fragmented in the
second quadrupole through incidental collision with neutral gas molecules. The characteristic product ion monitored in the third quadrupole is at m/z 201. The selected reaction
monitoring (SRM) chromatogram shows one peak with a retention time identical to
morphine. The final confirmation is based on the analysis of morphine standard.
Index
2D-GC, see two-dimensional gas
chromatography
AAS, see atomic absorption spectrometry
absorbance, 89–93, 95
specific, 90
test for, 364
see also Beer’s law
absorption maximum, 85, 92
absorption spectrum, 85
accuracy, 51–2, 317–18
assessment of, 318
acetaminophen, see paracetamol
acetic acid, 77–8, 393
acetone, 168
acetonitrile
as organic modifier, 152
as precipitant, 274, 278
acetylsalicylic acid (aspirin)
stability testing, 30
titration of, 74
UV spectrum of, 85
acid–base titrations, 67, 72–5
in aqueous environment, 389–93
in non-aqueous environment, 75–8, 393–6
color indicators for endpoint detection 73–4,
77
quantitative determination of raw materials
by, 389–96
acidity/alkalinity, 365
see also pH
acid value, 383
activated carbon, 150
active pharmaceutical ingredients (APIs)
assays of, 427–46
definition of, 1
patent protected, 19, 402
see also pharmaceutical raw materials
acylation, 207
ADME properties, 28
adsorption chromatography, see normal-phase
chromatography
AES, see atomic emission spectrometry
AFID, see alkali flame ionization detector
agarose, 265
alkali flame ionization detector (AFID), 204
alkylation, 207
amino acids, derivatization of, 207
p-amino-benzoic acid, 242
amiodarone hydrochloride, 370–371
amitriptyline
chemical characteristics of, 299–301
quantitiative determination in serum, 466–72
amitriptyline hydrochloride, 392
ammonia, 245–6
amphetamine
mass spectrum of, 237, 238, 245
retention on SPE column, 294–6
amu, see atomic mass units
analysis, 62
analyte, 33
analytical balance, 37–41
calibration of, 39–40
sensitivity of, 37–8
analytical measurement, 34
analytical procedure, 317
anhydrous acetic acid, 77–8, 393
anhydrous formic acid, 393
Introduction to Pharmaceutical Chemical Analysis, First Edition. Steen Honoré Hansen, Stig Pedersen-Bjergaard
and Knut Rasmussen.
2012 John Wiley & Sons, Ltd. Published 2012 by John Wiley & Sons, Ltd.
478
Index
anion exchangers, 293–4
anionic impurities, 372–6
anions, 211
anode, 211
anti-Stoke lines, 270–271
APCI, see atmospheric pressure chemical
ionization
APIs, see active pharmaceutical ingredients
apparent mobility, 215
appearance of solution, 361–3
applied field strength, 270
aspirin, see acetylsalicylic acid
assay, 337, 427
of diphenhydramine, 444–6
of doxapram, 432–3
of Fe2þ, 442–3
of fentanyl, 436–8
by HPLC, 429, 433–41
of hydrocortisone, 439–41
of omeprazole, 433–6
of paracetamol, 430–431
by titrations, 433, 442–6
by UV spectrophotometry, 427–33
see also quantitative determination
atmospheric pressure chemical ionization
(APCI), 247–8
atomic absorption spectrometry (AAS), 86–7,
115–22, 124–5
applications and performance, 121–2
instrumentation, 118–21
method development, 122
practical work, 122
atomic emission spectrometry (AES), 86–7,
123–5
instrumentation, 124
atomic mass units (amu), 233–4
atomization, 115
in a flame, 115, 117, 121, 122
in electrically heated furnaces, 121–2
atomization device, 120
autoprotolysis, 23–4
back-extraction, 286–7
barium chloride, 359, 361, 375
barium sulfate, 361, 375
base peak, 234
batch process, 8
beam chopping, 118
beclometasone dipropionate, 26–7
bed volume, 289
Beer’s law, 89–90, 92–3, 116, 120, 183
see also absorbance
benzene, 198
bioanalysis, 8, 453
in doping control, 457–8
in drug development, 453–5
in forensic and toxicological analysis, 456
in therapeutic drug monitoring, 455–6
see also bioanalytical method; biological
fluids
bioanalytical method, 460–466
calibration, 465–6
detection, 464–5, 472–6
liquid–liquid extraction, 463
protein precipitation, 274, 278–9, 462–3, 468
quantification, 465–6, 468–72
sample preparation, 461–4, 466–8
sampling, 460–461
separation, 464
solid-phase extraction, 463–4
biological fluids, 458–60
see also bioanalysis; bioanalytical method;
blood; plasma (part of blood); serum; urine
biological matrix, 458–60
blind samples, 62–3
blood
composition of, 458–9
sampling, 460–461
typical drug concentration in, 38
British Pharmacopoeia, 18, 330
monograph for paracetamol tablets, 409–12
bromazepam, mass spectrum of, 243
bromine
isotope patterns of, 241–2
release of, 243
Brownian movements, 175
“bubble-cell”, 215
buffers, 24, 152
phosphate, 152
running, 212
buffer system, 24
buoyancy equation, 40
burettes, 46, 47, 66
byproducts, 328–9
C18 materials, 148, 149, 150, 151, 291, 468
CAD, see charged aerosol detector
caffeine, determination of, 91
calcitriol, identification of, 358
calcium sulfate, 167
Index
calibration
of analytical balance, 39–40
of analytical methods, 49–50
in bioanalysis, 465
definition of, 50
direct, 116
of pipettes, 43–4
see also calibration curve; calibration methods
calibration curve, 49
calculation of, 59
see also calibration
calibration methods, 310–314
external standard, 310–312, 465
internal standard, 313–14, 315, 465
normalization, 314, 317
one-point calibration, 312
standard addition, 51, 92, 117, 314, 316
see also calibration
calomel electrode, 69–70
capillaries, 213, 217–18, 225–9
coated, 225–9
silica, 213, 218
capillary columns, 140, 198–9
injection systems for, 200–202
capillary electrochromatography (CEC), 224
capillary electrophoresis (CE), 199, 211–29
capillaries, 213, 217–18, 225–9
chiral separations in, 224–5
electroosmotic flow, 213–14
instrumentation, 214–17
non-aqueous, 229
principle of, 211–12
sample introduction, 218–20
capillary supercritical fluid
chromatography, 265
capillary zone electrophoresis (CZE), 214,
221–2
chiral separations in, 224–5
capsules, monograph for, 406–8
carbon
isotopes of, 236, 240
number of atoms, 240–241, 242
carbon dioxide, 264
Carbowax, 196
carboxylic acids, 293–4
carrier gas, 191, 192–3, 195–6
carrier support, 199
castor oil, monograph of, 337, 338
catecholamines, oxidation of, 186
cathode, 211
479
cation exchangers, 293
cationic impurities, 376–8
see also heavy metals
cations, 211
CE, see capillary electrophoresis
CEC, see capillary electrochromatography
cellulose fibers, 263
centrifuge tube, 279
cerium sulfate, 81
charged aerosol detector (CAD), 187
chemical interference, 122
chemical ionization (CI), 244–6
chemical reference substance (CRS), 50, 346
chiral center, 353
chirality
introducing into chromatographic
system, 157–8
visualization of concept, 354
chiral mobile phases, 157–8
chiral selectors, 225
chiral separations
in capillary electrophoresis, 224–5
in chromatographic systems, 156–8
chiral stationary phases, 157–8
chlorambucil
fragmentation, 231–2
ionization, 231
mass spectrum of, 233, 234, 235, 236
chlorcyclizine hydrochloride, identification
of, 361, 362
chlorides
identification of, 359, 361
test for, 372–6
chlorine
isotopes of, 235, 241, 242
release of, 243–4
chloroform, test for, 373–4
chromatogram, 129, 140
chromatographic development, 168–9
chromatographic separation principles, 141–61
chiral separations, 156–8
hydrophilic interaction chromatography
(HILIC), 156
ion exchange chromatography, 160–161
normal-phase chromatography, 140, 142–7
reversed-phase chromatography, 140,
148–56
size exclusion chromatography (SEC),
158–60
see also chromatography
480
Index
chromatographic separation process, 175–7
see also chromatography
chromatography, 127–40
column, 141
column efficiency in, 133–5
general principles, 127–31
overview of techniques, 140
peak symmetry, 136–7
resolution, 138–9
retention, 131–3
selectivity, 130–131, 135–6
see also chromatographic separation
principles; chromatographic separation
process; gas chromatography (GC); high
performance liquid chromatography
(HPLC); thin-layer chromatography (TLC)
chromophores, 95
clinical development, 454
clonazepam, 243–4
cis-clopenthixol, 29
13
C NMR spectroscopy, 268
cocaine, 256
Code, see World Anti-Doping Code
codeine, 221–2
chemical characteristics of, 301–2
UV spectrum, 94
cold trap, 201–2
color indicators, 69
for acid–base titrations, 73, 77
for non-aqueous titrations, 77
for redox titrations, 80–81
color reactions, 344, 403, 412
column chromatography, 141
see also gas chromatography (GC); liquid
chromatography (LC)
column efficiency, 133–5, 140
columns, 127
capillary, 140, 198–9, 200–202
in gas chromatography, 198–9
in high performance liquid chromatography
(HPLC), 177–80
packed, 199, 202
solid-phase extraction, 288–9
column void volume, 131
combined glass electrode, 69, 71
complex formation titrations, see
complexometric titrations
complexometric titrations, 67, 81
concentration, 35–7
expressing as percentage, 36–7
of drug in human blood, 38
concentration zone, 166
conditioning, of SPE column, 288, 289
confidence intervals, 54–5
constitutional isomers, 28
containers, 406–8
continuous spectrum, 86
conversion
from mg/ml to M, 36
from mg/ml to mg/ml, 36
from M to mM, 35
from ng/ml to nM, 38
cool on-column injection, 200, 202
Coomassie brilliant blue dye, 266
corona discharge detector, 186
correction factor, 43
counter ions, 77, 155
titration of, 77
creatinine, 460
critical point, 264
CRS, see chemical reference substance
cyclodextrins, 225, 226
cyclohexane, 282
CZE, see capillary zone electrophoresis
DAD, see diode array detector
Daltons, 234
degassing, 188
degree of freedom, 53, 54
densitometry, 171
deoxyribonucleic acid (DNA), 265, 266
derivatization, 206–7
of amino acids, 207
of enantiomers, 157, 225
N-desmethylclobazam, 255, 256, 257
detection limit, 320–321
detectors
alkali flame ionization (AFID), 204
capillary electrophoresis, 215–16
charged aerosol (CAD), 187
combination of, 187
corona discharge, 186
diode array, 183–4
electrochemical, 182, 186
electron capture (ECD), 204–6
evaporative light scattering (ELSD), 186–7
flame ionization (FID), 203
fluorescence, 182, 184–6, 216
gas chromatography, 203–206
general, 182
Index
HPLC, 182–7
mass spectrometry, 206
nitrogen–phosphorus (NPD), 203–4
refractive index, 182, 186
specific, 182
thermal conductivity (TDC), 204, 205
thermionic, (TID), 204
universal, 186–7
UV, 182–4, 189, 215
determination, 33–4, 62
see also quantitative determination
deuterium lamps, 96, 183
dextrorotatory compounds, 355
diamond, 108, 150
diastereomers, 28, 157
diatomaceous earth, 166, 199
diazepam, 417–19
dichloromethane, 282, 283
diethyl ether, 283
diffuse reflection, 110–111, 112, 113
diffusion, 175–5
longitudinal, 175, 214
diffusion coefficient, 215
digits, 58–60
see also significant figures
dilution equation, 48–9
dilutions, 47–9
diode array detector (DAD), 183–4
diphenhydramine, 444–6
dipole–dipole interactions, 143, 197, 297
direct calibration, 116
dispersion interactions, 143
dissociation constant, 280
dissolution, 451
distribution constant, 25, 128, 132, 144
see also partition ratio
distribution ratio, 26, 280–281, 283
divinylbenzene, 291
DNA (deoxyribonucleic acid), 265,
266
dopamine hydrochloride, 365
doping control analysis, 457–8
dosage, adjusting, 8, 455–6
dosage forms, 2–3
general monographs for, 406–9
dose–response relationship, 45
double-beam atomic absorption
spectrometer, 121
double-beam IR spectrophotometer, 107
double-beam UV spectrophotometer, 98–9
481
double focusing mass spectrometry, 252–3
doxapram, 432–3
drug development process, 453–5
drug discovery, 453–4
drugs, see final pharmaceutical products; medical
products
drying gas, 246, 251
ECD, see electron capture detector
EDTA, see ethylenediaminetetraacetic acid
effective field strength, 269–70
EFPIA, see European Federation of
Pharmaceutical Industries and
Associations
EI, see electron ionization
electrochemical detector, 182, 186
electrodes, 69–71
electrodispersion, 214
electrokinetic injection, 220
electromagnetic radiation, 83–8
and atoms, 86–8
and molecules, 85–6
monochromatic, 83, 84, 89
types of, 84
electromigration, 220
electron capture detector (ECD), 204–6
electron ionization (EI), 236–45
differentiated from electrospray ionization
(ESI), 246–7
electroosmosis, 213
electroosmotic flow (EOF), 213–15, 225–7
electropherogram, 215, 216
electrophoretic migration, 214–15
electrophoretic mobility, 211
electrospray ionization (ESI), 246–7
ELSD, see evaporative light scattering
detector
eluate, 275, 288
eluent, 288
elution, 275, 288, 292
EMA, see European Medicines Agency
emission, 88
emission wavelength, 184
enantiomers, 28–9
derivatization to diastereomers, 157, 225
optical rotation, 353–6
separation of, 156–8
endcapping, 149
enrichment, 277–8
EOF, see electroosmotic flow
482
Index
ephedrine
determination of water in, 381–2
quantitative determination by acid–base
titration, 392
UV spectrum of, 94
ephedrine hydrochloride, 394, 396
equivalence point, 65–6, 69
errors
random, 50–51
systematic, 50–51
titration, 68
ESI, see electrospray ionization
esomeprazole magnesium trihydrate, 364
essential oils, 208, 386–9
ethanol, 198
Ethidium bromide, 266
ethyl acetate, 282, 283
ethylenediaminetetraacetic acid (EDTA), 81
EU, see European Union
European Federation of Pharmaceutical
Industries and Associations (EFPIA), 4
European Medicines Agency (EMA), 14
European Pharmacopoeia (Ph.Eur.), 18–19
contents of, 18, 331–7
general monographs for dosage forms, 406–9
monograph for capsules, 406–8
monograph for castor oil, 337, 338
monograph for paracetamol, 332–4
standards and tests for raw
materials, 330–337
updates, 402–3
validation of test methods in, 325
European Union (EU) 14–15
evaporative light scattering detector
(ELSD), 186–7
exact mass, 234
excipients
definition 2
functions of, 3
of paracetamol syrup, 3
of paracetamol tablet, 3
see also pharmaceutical raw materials
excitation wavelength, 184
excited state, 86
exclusion volume, 158
external standard calibration, 310–312, 465
extract, 275
extraction solvents, 279, 282–3, 287
fatty acids, 383, 389
analysis by gas chromatography (GC), 208,
386
methylation in oils, 207
fatty oils, 330, 337, 385
acid value for, 383
analysis by gas chromatography (GC), 386,
389
unsaponifiable matter in, 386
FDA, see Food and Drug Administration
Fe2þ, 442–3
fencamfamin, 239
fenoterol hydrobromide, 366
fentanyl
assay of, 436–8
identification of, 424–5
ferrous fumarate, 397, 442–3
FID, see flame ionization detector
final pharmaceutical products
assay by HPLC, 429, 433–41
assay by titrations, 433, 442–6
assay by UV spectrophotometry, 427–33
chemical testing, 406–12, 446–51
dissolution test, 451
identification of APIs by HPLC, 417, 422–5
identification of APIs by IR
spectrophotometry, 412, 413–17
identification of APIs by thin-layer
chromatography, 425, 428–9
identification of APIs by UV
spectrophotometry, 412, 417–22
microbiological testing, 406
pharmaceutical testing, 406
quality control of, 405–6
test for related substances, 446–9
test for uniformity of content, 449–50
first identification, 335, 344, 345
flame emission spectroscopy, 123, 124
flame ionization detector (FID), 203
flame photometry, 123–4
flow rate, 175
flow spoilers, 120
fluorescence, 86
fluorescence detector, 182, 184–6, 216
fluorescence intensity, 185–6
fluoride-selective electrode, 263
fluoxetine
identification by HPLC, 422–4
identification by IR
spectrophotometry, 414–15
fluoxetine hydrochloride, 414, 422
Index
flupentixol decanoate
identification by TLC, 428
identification by UV
spectrophotometry, 419–20
Food and Drug Administration (FDA), 14
foreign anions, see anionic impurities
foreign cations, see cationic impurities
forensic analysis, 456
formic acid, 152, 167–8, 396
anhydrous, 393
formulation, 2–3
Fourier transformation, 108
Fourier transform IR (FTIR)
spectrophotometers, 108
fragmentation, 231–2, 240, 241
fragment ions, 231, 237, 240
free atoms, 88, 123
frequency, 83–4
full scan, 253–4
furosemide
chemical characteristics of, 304–5
test for chlorides and sulfates in, 375–6
gas chromatography (GC), 140, 141, 191–209
analysis of essential oils by, 386–8, 389
analysis of fatty oils by, 386, 389
apparatus, 192–3
applications, 208–9
carrier gas, 191, 192–3, 195–6
columns, 198–9
derivatization, 206–7
detectors, 203–6
injection systems, 200–202
mobile phases for, 191
outline of the principle, 192
selectivity, 197–8
static headspace, 372
stationary phases for, 191, 193, 196–8
temperature, 193–4
test for residual solvents by, 372, 373–4
two-dimensional, 209
see also chromatography
gas chromatography–mass spectrometry (GCMS), 206, 231, 236–7, 250, 253
applications, 257
for identification of unknown substances in
biological fluids, 209, 473–5
for quantitative analysis, 256–7, 471–2
gas–liquid chromatography (GLC), 193
gas–solid chromatography (GSC), 193
483
Gaussian distribution, 52
Gaussian peaks, 129, 133
GC, see gas chromatography
GC-MS, see gas chromatography–mass
spectrometry
gel electrophoresis, 265–7
two-dimensional 268
gel filtration chromatography, 158
gel permeation chromatography (GPC), 158
generic drugs, 4
glass electrode, 69, 70–71
GLC, see gas–liquid chromatography
Globar source, 108
GLP, see Good Laboratory Practice
GMP, see Good Manufacturing Practice
Good Laboratory Practice (GLP), 16
Good Manufacturing Practice (GMP), 10, 16,
20, 21
GPC, see gel permeation chromatography
gradient elution, 181–2
ground state, 86
GSC, see gas–solid chromatography
gypsum (calcium sulfate), 167
HCl, 68, 72, 73
headspace injection, 208
heavy metals, 343
test for determination of, 376–8
height equivalent to a theoretical plate
(HETP), 175, 178, 195
helium, 195, 388
hematocrit, 458
hemolysis, 459
Henderson–Hasselbalch equation, 24, 26
heptafluorobutyrate, 155, 156
heptane, 197, 282
herbal medical products, 2
Hertz, 83
HETP, see height equivalent to a theoretical plate
hexane, 283
high performance liquid chromatography
(HPLC), 140, 173–90
columns, 177–80
detectors, 182–7
injectors, 187–8
main components of the system, 173–4
mobile phases, 188–9
pumps, 174, 180–181
reporting results, 189–90
solvents for sample preparation, 189
484
Index
high performance liquid chromatography
(HPLC) (Continued)
system suitability testing (SST), 325, 399
use for assays, 429, 433–41
use for bioanalysis, 464
use for identification, 356–9, 417, 422–5
use for quantitative determination, 396–400
use in test for related substances, 367–70,
446–9
van Deemter equation for, 176
see also chromatography
high performance thin-layer chromatography
(HPTLC), 166, 168, 169, 171
high-resolution mass spectrometry, 248–9
high throughput screening systems (HTS), 454
HILIC, see hydrophilic interaction
chromatography
1
H NMR spectroscopy, 268, 269
hollow-cathode lamp, 118–20
horizontal development, 168–9
HPLC, see high performance liquid
chromatography
HPLC-UV, 101
HPTLC, see high performance thin-layer
chromatography
HTS, see high throughput screening systems
hydrocarbons, 144–5, 146, 282, 282
hydrochlorides, 302, 359, 390
hydrocortisone
assay by HPLC, 439–41
chemical characteristics of, 307–8
quantitative determination by UV
spectrophotometry, 401–2
hydrodynamic injection, 218–20
hydrogen, 195
bo
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )