RESEARCH PAPER Vibration isolation system for cryocoolers on the XRISM Susumu Yasuda ,a,* Yoh Takei ,a Atsushi Okamoto,a Keiichi Yanagase ,a Kazuki Watanabe,b and Seiji Yoshidac a ABSTRACT. Japan Aerospace Exploration Agency, Tsukuba, Japan b WEL Research, Ichihara, Japan c Sumitomo Heavy Industries, Ltd., Niihama, Japan We discuss the design and performance of the vibration isolation system (VIS) used for the Resolve microcalorimeter on the X-Ray Imaging and Spectroscopy Mission (XRISM) spacecraft. Resolve makes high spectral resolution observations of celestial X-rays. XRISM was conceived as a recovery mission following the short-lived Hitomi mission. As such, the design of Resolve is similar to that of the soft X-ray spectrometer (SXS). A VIS was initially developed for the Hitomi SXS after discovering that vibrations from the cryocooler degraded the sensor’s performance. However, the VIS for Resolve was completely redesigned to include a hold-andrelease mechanism to avoid damage to the compressor during launch or ground-based sine-vibe tests. The XRISM spacecraft was successfully launched in September 2023, confirming the functionality of the VIS while in orbit. © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI. [DOI: 10.1117/1.JATIS.11.4.042004] Keywords: X-ray observation; microcalorimeter; cryocooler; vibration isolation Paper 24213SS received Dec. 13, 2024; revised Feb. 13, 2025; accepted Feb. 21, 2025; published Mar. 12, 2025. 1 Introduction Cryogenic techniques are often used in space missions that require extremely high sensitivity or low noise measurements. Cryogenic instruments have revealed an unprecedented view of the universe at various wavelengths. Examples include COBE,1 WMAP,2 and Planck3 for observing the cosmic microwave background; AKARI,4 Spitzer,5 Herschel,6 and JWST7 for observing the universe in the infrared; and Hitomi8 for X-rays. Cryogens, particularly liquid (or superfluid) helium, have been used for a long time in space missions. In addition, cryocoolers have recently been used to reduce the heat load on the cryogen (e.g., AKARI, Hitomi, and XRISM), or even as a primary cooling method (e.g., JWST). Furthermore, many planned missions (e.g., LiteBIRD,9 Athena X-IFU,10 Line Emission Mapper,11 Probe Far-Infrared Mission for Astrophysics,12 and Black Hole Explorer13) will operate without cryogen, instead relying on cryocoolers, so that the mission lifetime is not limited by the amount of cryogen. Cryocoolers [e.g., Stirling coolers (ST), pulse-tube coolers, and Joule-Thomson coolers (JT)] utilize repeated cycles of compression and expansion of the working gas for cooling. The mechanical movement of compressors (CMPs) and other parts, along with the flowing working gas, creates vibration, often disturbing highly sensitive onboard instruments. Mitigating these vibrations is therefore important. *Address all correspondence to Susumu Yasuda, yasuda.susumu@jaxa.jp J. Astron. Telesc. Instrum. Syst. 042004-1 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM This paper presents the design and performance of a vibration isolation system (VIS) for the Resolve instrument14 on the X-ray imaging and spectroscopy mission (XRISM).15 Resolve is a microcalorimeter-based instrument enabling observations of celestial X-rays with high spectral resolution (E∕ΔE ∼ 1000, or <7 eV full-width half-maximum at 6 keV). As XRISM was a recovery mission for the short-lived Hitomi, the design of Resolve is fundamentally similar to the soft X-ray spectrometer (SXS),16 the microcalorimeter instrument onboard Hitomi. Thus, the design of Resolve’s cooling chain is the same as that of SXS. Details can be found in Refs. 17 and 18. The cross-section of the Dewar (DWR) and its cryogenic architecture are shown in Figs. 1 and 2, respectively. The detector is installed inside the helium DWR, a vacuum vessel containing a helium tank. A three-stage adiabatic demagnetization refrigerator (i.e., ADR19) achieves the operating temperature of the detector, 50 mK. The assembly containing the detector and the ADR (called the calorimeter spectrometer insert; CSI) is mounted on a 39 L helium tank and cooled to around 1 K by superfluid helium. There are multiple radiation shields between the helium tank and the DWR outer shell [i.e., the DWR main shell (DMS)]: the JT shield, inner vapor-cooled shield (IVCS), middle vapor-cooled shield, and outer vapor-cooled shield (OVCS). The radiation shields are cooled by the evaporated helium flowing through the vent-line plumbing. Furthermore, the JT shield is cooled by a He-4 JT cooler, whereas the IVCS and the OVCS are cooled by two double-stage Stirling coolers (2ST coolers). The first stage of the two 2ST coolers cools the OVCS, whereas the second stage cools the IVCS. These 2ST coolers are shield coolers (SC). The other two 2ST coolers cool the inner DWR plumbing of the JT cooler. These 2ST coolers are called pre-coolers (PC). In total, four 2ST coolers are used in the Resolve cooling chain. The CMPs and cold heads (CH) of the 2ST coolers and the CMPs of the JT coolers are mounted on the DMS. Hitomi SXS development revealed that the vibration of the cryocoolers affects the detector’s temperature stability and spectroscopic performance. The CHs and JT cooler produce only cyclic vibration, so it was possible to avoid sensitive frequencies of the calorimeter by changing their driving frequency. However, the CMPs produce not only cyclic vibration but also non-cyclic or random vibration due to the ball bearings supporting the movement of the pistons. Therefore, a Fig. 1 Cross-section of the Resolve Dewar.14 J. Astron. Telesc. Instrum. Syst. 042004-2 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Fig. 2 Cryogenic architecture of the Resolve Dewar. VIS was added between each CMP and the DMS to mitigate this, decoupling the mechanical vibration of the 2ST CMPs from the DMS.20 This improved the energy resolution. However, the mechanical load applied to the 2ST CMPs during launch was only marginally within limits because the VIS amplified the acceleration below 100 Hz. For XRISM, it was decided to install VIS units for 2ST CMPs but not for JT CMPs and 2ST CHs, as was done for Hitomi. The design of the VIS for Resolve incorporates a hold-and-release mechanism (HRM) to prevent damage to the CMPs during launch or sine-vibe tests on the ground. The VIS is an integrated system with a vibration isolator, thermal path, and hold-and-release functions. The XRISM spacecraft was launched in September 2023, and the functions of the VIS have been verified in orbit. The 2ST and JT coolers installed in XRISM were developed by Sumitomo Heavy Industries Ltd (SHI). To reduce wear on the lubrication material of the piston surfaces, ball bearings were introduced as a guiding mechanism for the pistons. Although flexure springs were also considered for this purpose, ball bearings were selected for the compact design of the coolers. The first 2ST cooler featuring ball bearings was used in the cooling system of AKARI,4 which was launched in 2006. At that time, vibration from the ball bearings was not identified as an issue. Thus, based on the successful flight heritage of AKARI, the same 2ST design was adopted for Hitomi. However, during the development of Hitomi, it became clear that vibration from the ball bearings was problematic. Due to time constraints, a complete redesign of the 2ST was not feasible. Instead, countermeasures implemented by VIS proved successful, and the same approach was applied to XRISM. The JT cooler with ball bearings was also utilized in JEM/SMILES,21 which was launched in 2009. However, due to issues related to impurity generation, a version featuring flexure springs was developed and adopted for Hitomi. Currently, SHI is working on a 2ST cooler that incorporates flexure springs to minimize microvibration.22 2 VIS Requirements The VISs are positioned between the DMS and four CMPs (SC-A, SC-B, PC-A, and PC-B). The HRM, introduced for Resolve, has two states: released and locked. In the released state, the VIS minimizes vibration transmission from the CMPs to the DWR, ensuring that the sensitive detector inside the DWR experiences negligible mechanical vibration. On the other hand, in the locked J. Astron. Telesc. Instrum. Syst. 042004-3 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM state of the HRM, the VIS can withstand the harsh mechanical environment during launch and ground vibration testing without amplifying mechanical loads on the CMPs. In addition, the VIS facilitates heat transfer from the CMPs to the DMS or a cooling plate, aiding in cooling the CMPs during operation. In this section, essential requirements for the VIS are described. 2.1 Performance and Functional Requirements 2.1.1 Lifetime As the XRISM satellite’s design lifetime is 3 years, the VIS must maintain its performance requirements in orbit for at least 3 years. In addition, the VIS should withstand up to 3 years of storage on the ground (under 1 G and in a normal atmosphere) before launch. 2.1.2 Vibration transmissibility The primary characteristic of the VIS is its vibration transmissibility (τ), which is defined as the ratio of the output force/moment (Fvo ) to the input force/moment (Fvi ). The definitions of these forces and moments are shown in Fig. 3. A lower transmissibility indicates superior vibration isolation. The requirement for vibration transmissibility is the same as that for Hitomi. When the micro-disturbances due to the cryocoolers were studied for Hitomi, the vibration level threshold that produces the undesired temperature disturbance (thus causing detector performance degradation) was measured using the Hitomi DWR engineering model in which the CSI engineering model was installed.23 Given the similarity of the CSI and the DWR between Hitomi and XRISM, the same threshold level was assumed for XRISM. When the threshold was determined, a six-degree of freedom (DOF) vibrator was used to simulate the CMP vibration, and the allowable level within a specified frequency range was determined. The maximum permissible transmissibility was calculated as the quotient of the allowable vibration level and the force acting on the CMP. Due to the performance characteristics of the vibrators used, the defined frequency range was limited to 10 to 600 Hz for Fx and Fz and 100 to 600 Hz for the other axes. The requirement in the low frequencies of FY, MX, MY, and MZ, which were not determined experimentally, was set by extrapolation. The transmissibility of the VIS must not exceed the thresholds represented by the blue lines in Fig. 4. We empirically knew that the detector was not significantly affected by vibrations below 100 Hz. As shown in Fig. 4, the required transmissibility for Fx and Fz in the frequency range of 10 to 100 Hz was found to be ∼20 to 50. The passive isolator amplifies vibration most at its resonant frequency, with the maximum value determined by the Q factor. As it was evident that the resonant frequency of the isolator would be below 25 Hz and the Q factor would be below 10, we assumed the extrapolated requirement is acceptable for frequencies below 100 Hz, other than for Fx and Fz. Our assumption was finally verified by the verification test using the detector, described in Sec. 4.6.2. This requirement must be met both in orbit and in 1 G on the ground, even when the thermal straps (TSs) and the cryocooler’s capillary tubes are attached. However, this requirement does not apply in the locked state of the HRM. CMP z x y CG of CMP x z y VIS origin of Fvi DWR origin of Fvo W=Fvo/Fvi Fig. 3 Definition of vibration transmissibility (τ). J. Astron. Telesc. Instrum. Syst. 042004-4 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Fig. 4 Requirement of VIS transmissibility. 2.1.3 Thermal conductivity The VIS must be able to transfer a specified amount of heat from the CMPs for cooling. There are two distinct thermal conductivity requirements that correspond to the different CMP heat transfer paths. For SC-A, SC-B, and PC-A, with heat dissipated through the DWR, the thermal conductance between the CMP and the DMS should be at least 2 W∕K, with each CMP able to transfer more than 63 W of heat. For PC-B, where the heat is dissipated to the cold plate cooled by a spacecraft heat pipe, the thermal conductance between the CMP and the cold plate should be equal to or exceed 1 W∕K. 2.1.4 Compressor load at vibration test To prevent damage to a CMP mounted on the VIS, the maximum acceleration applied to the CMP must not exceed the specified level for the CMP under acceptance test (AT) conditions. These conditions include sine vibration, acoustics, shock, and quasi-static load. In other words, the range of the acceleration amplification is specified. This requirement is related to the stiffness requirements of the VIS in the locked state. If the stiffness is insufficient, the vibration transmitted to the CMP could be amplified, causing damage. This applies only when the HRM is locked as the HRM is supposed to be locked whenever mechanical loads are applied. 2.1.5 Release operation requirements The shock level generated by the HRM release must not exceed the CMP’s acceptance level. The shock is also transmitted to the DWR, where the most sensitive component is a thin filter located at the top of the DMS. Consequently, the allowable shock level at the top of the DWR is defined as 100 Gsrs within the frequency range of 1000 to 4000 Hz. During the release operation, the displacements of CMPs must remain within a specified range. When the HRM is released, the CMP moves from the locked position to the released position. Each CMP is connected to the corresponding CH via a capillary tube that facilitates the flow of helium, the working fluid. In addition, SC-A and SC-B are equipped with loop heat pipes (LHPs) to transfer heat to the spacecraft’s two radiator panels. Both the capillary tube and the LHP are constructed from stainless steel and have maximum displacement requirements to avoid plastic deformation. The allowable displacements are defined at the attachment points of the capillary tube and the LHP, as shown in Table 1. This requirement must be met under both 1 G conditions and in orbit. J. Astron. Telesc. Instrum. Syst. 042004-5 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 1 Requirement on the maximum allowed CMP displacement during release. Capillary attachment point LHP attachment point Item Y (mm) Z (mm) Y (mm) Z (mm) SC-A ±7.8 ±6.7 ±5.0 ±1.0 SC-B ±7.8 ±6.7 ±5.0 ±1.0 PC-A ±1.9 ±3.4 N/A N/A PC-B ±3.4 ±8.3 N/A N/A 2.2 Environmental Requirements The VIS must withstand the mechanical loads during launch and ground testing (sine vibration, acoustics, shock, quasi-static loads). The HRM was introduced to fulfill this requirement. Fatigue-induced damage that can occur both on the ground and in orbit must be considered. Three sets of temperature ranges are defined for the VIS’s thermal environment requirements. The storage temperature range, within which no damage occurs, is defined as −60 to þ 40°C. In addition, the performance assurance temperature range, within which the VIS operates properly, is defined as −30 to þ 40°C. Finally, the temperature range during launch and ground testing is defined as þ 10 to þ 40°C. The VIS must be compatible with the in-orbit environment in terms of vacuum, radiation, atomic oxygen, and UV radiation. The VIS must function under 1 G gravity (on the ground) and zero gravity (in orbit). In particular, the vibration transmissibility requirement must be satisfied, and the pin-puller release must operate without interference under both conditions. 2.3 Interface Requirements The VIS must fit within the volume provided by the DWR design. Although this may seem trivial, it imposes stringent constraints on the VIS design because the DWR design was established for the Hitomi SXS, and only minor modifications were possible. The HRM pin-puller should be activated by Hitomi compatible electronics in orbit as no changes to the electronics design are preferred. Therefore, the pin-puller should have a resistance of 16.0 1.0 Ω (including the harness on the DWR) and must be released with a current pulse ranging from 1.0 to 1.6 A for 30.0 1.0 ms. This is required for the in-orbit release operation of the pin-puller as a spacecraft component applies the pulse. However, this requirement does not apply when the HRM operates with ground support equipment (GSE). 2.4 Operational Requirements On the ground, during the microcalorimeter performance test, the HRM must be released to meet the transmissibility requirement. However, it must be locked during the vibration test to withstand acceleration. The HRM should be capable of locking and releasing at least 20 times during ground testing without disassembling the satellite. Once the VISs are installed in the satellite, they cannot be accessed manually; therefore, the HRM must be capable of being locked remotely while on the ground. Meeting the thermal conductivity requirement during ground testing (except during the thermal-vac test) is unnecessary as spot coolers can effectively cool CMPs under atmospheric conditions. In orbit, the VIS HRM should be released only once upon command from the ground. The cryocoolers need to be activated as soon as possible after launch to preserve the cryogen (superfluid helium). Therefore, the HRM release operation should be executed immediately after launch as it must occur before turning on the cryocoolers. J. Astron. Telesc. Instrum. Syst. 042004-6 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM CMP CMP plate TS-type C Hold release mechanism (HRM) (remotely resettable pin-puller) TS-type B TS-type C DWR plate Isolator strut (-Y) Isolator strut (+Y) TS-type B HRM reset mechanism Isolator strut (Z) Fig. 5 VIS (SC-A, PC-A) exploded view. 3 Design of VIS 3.1 VIS Design Overview Figure 5 is an exploded view of the VIS. The VIS comprises a DWR plate, a CMP plate, six isolator struts, and TSs. The DWR plate incorporates an HRM release mechanism and a reset mechanism. The CMPs are attached to this CMP plate. During ground performance tests or in orbit after unlocking the HRM, the CMP is supported by the six isolator struts. However, the HRM holds the CMP plate in place during launch or ground vibration tests. The HRM reset mechanism, a unique feature of the XRISM VIS, remotely resets the previously released HRM. Figure 6 shows the design variation of the VIS. The design of the VIS for SC-A and PC-A is identical, whereas the design for SC-B is a mirror image of that for SC-A. For SC-A, SC-B, and PC-A, four TSs connect the CMP and the DWR plate, whereas for PC-B, a single TS connects the CMP to the cooling plate on the satellite. 3.2 Vibration Isolator Strut Figure 7 shows the configuration of the vibration isolator struts. The CMP is supported over the DWR plate by three pairs of vibration isolator struts (also referred to as isolator bipods). Figure 8 shows the structure of a single vibration isolator strut, comprising an upper component, a housing, two compression coil springs, and six-leaf springs made of single crystal shape memory alloy (SCSMA). Fig. 6 Three VIS designs: (a) for SC-A and PC-A, (b) for SC-B, and (c) for PC-B. J. Astron. Telesc. Instrum. Syst. 042004-7 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Attachment to CMP Flexible joint Joint rod Flexible joint Attachment to DWR plate Fig. 7 Configuration of vibration isolator struts. Stage Flexible joint SCSMA leaf spring SCSMA leaf spring Joint rod Compression coil spring Center shaft Upper component Housing Stage Stage Center shaft Stage Compression coil spring Housing Flexible joint (a) (b) Fig. 8 Structure of vibration isolator strut: (a) perspective view, (b) conceptual diagram. The compression coil springs and SCSMA leaf springs connect the upper component to the housing. One end of the isolator strut is attached to the DWR plate using a flexible joint, whereas the other end is connected to the CMP via a joint rod and a flexible joint. Initially, an attempt was made to use ball joints to connect the struts; however, issues such as lubricant wear, particle generation, and frictional instability were presented by the ball joints. Consequently, flexible joints were opted for over ball joints. Figure 9(a) shows the design of the flexible joint, which is made from A-286 stainless steel. We modeled the flexible joint as a spring element using static analysis [Fig. 9(b)], and Table 2 shows the results. Using this modeled spring element, we conducted a modal analysis of the VIS in the unlocked state (Fig. 10). In this model, the CMP was modeled as a mass. The results are shown in Table 3. The label “pin joint” refers to using a three DOF constraint to connect the struts, and “flexible joint” refers to using the spring element to connect them. As shown, the Fig. 9 (a) Design of flexible joint. (b) Static analysis model of a flexible joint. J. Astron. Telesc. Instrum. Syst. 042004-8 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 2 Stiffnesses of the designed flexible joint. Lateral stiffness (N/mm) 100 Axial stiffness (N/mm) 352 Bending stiffness (Nm/deg) 1.2 Torsion stiffness (Nm/deg) 24.4 Z X Y Fig. 10 Modal analysis model of VIS in an unlocked state. Table 3 Result of modal analysis in unlocked state. Frequency (Hz) Mode Pin joint Flexible joint Modal shape 1 4.3 3.5 Y rotation 2 6.8 6.6 X rotation 3 8.1 7.3 Z translation 4 10.2 9.6 X translation 5 11.5 10.3 Z rotation 6 11.6 11 Y translation flexible joint lowers the modal frequencies compared with the ideal pin joint. This reduction is due to the axial stiffness of the flexible joint; however, its effect was not critical. The stiffness and natural length of the isolator strut can be adjusted by the two compression coil springs. The natural frequencies of the suspended CMP were designed to be ∼4 to 12 Hz in six degrees of freedom, thus avoiding the CMP’s drive frequency of 15 Hz. Furthermore, the static displacements of the strut under 0 G and 1 G conditions are also important considerations. Taking these conditions into account, the parameters of the compression springs were designed. These struts have a simple lightweight structure, entirely made of metal materials without using polymers such as rubber or oil. As a result, they exhibit extremely low outgassing and aging. 3.2.1 SCSMA leaf spring The SCSMA leaf springs act as damping elements.24–27 Shape-memory alloys (SMAs), widely utilized in practical devices, exhibit remarkable properties such as shape-memory recovery and J. Astron. Telesc. Instrum. Syst. 042004-9 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Stress, MPa 300 200 60 C 40 C 20 C 100 0 0 2 4 6 Elongation, % 8 10 Fig. 11 Stress–strain curve of SCSMA.28 4.0 mm R= 9.5 mm 50.0 mm T=0.8 mm Fig. 12 Detailed design of SCSMA leaf spring. superelasticity. Although typical SMAs are polycrystalline, the CuAl-based single-crystal shapememory alloy (SCSMA) has recently been developed.28 SCSMA possesses superior shapememory recovery and superelasticity. Superelasticity results from a stress-induced conversion from austenite to martensite when stress exceeds a critical level and reversion from martensite to austenite when stress falls below another critical level. These phenomena produce a pair of constant stress plateaus in the stress–strain plot at specific temperatures. Single crystal superelasticity is characterized by an abrupt change in the slope of the stress–strain plot at a combination of the alloy’s stress, strain, and temperature characteristics. A typical stress–strain property of SCSMA is shown in Fig. 11. For an elongation between roughly 2% and 8%, the stress–strain curve exhibits hysteresis, which is used for damping. Figure 12 is a detailed design of the SCSMA leaf spring. The thickness of the leaf spring (T ) and curvature radius (R) in the neutral position is 0.8 and 9.5 mm, respectively. Therefore, the maximum strain of the leaf spring is T ∕2R ¼ 4.2% at which SCSMA shows significant hysteresis. 3.2.2 Characteristics of isolator strut A characteristic acquisition test was conducted on the breadboard model (BBM) of the isolator strut. A displacement–force curve was obtained using a fatigue testing machine (Fig. 13). An example of the displacement–force curve obtained in the test is shown in Fig. 14(a). The stiffness (ks ) and the loss factor (ηs ¼ tan δ) are calculated as follows: ks ¼ EQ-TARGET;temp:intralink-;e001;114;154 Fs1 ; Xs ηs ¼ tan δ ¼ EQ-TARGET;temp:intralink-;e002;114;107 Fs0 : Fs1 (1) (2) Here, Xs , Fs1 , and Fs0 represent the maximum displacement, the force at the maximum displacement, and the force at zero displacement, respectively. The displacement dependency J. Astron. Telesc. Instrum. Syst. 042004-10 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Fig. 13 Test configuration: (a) Isolator strut. (b) Fatigue testing machine. (c) Constant temperature chamber. Xs (b) 20 4 N/mm 0 10 0.1 5 –4 –8 –1.5 0 –1 –0.5 0 0.5 1 0 0 1.5 0.2 Displacement (mm) 1 mm 0.5 mm 0.1 mm 0.05 mm 0.6 0.8 1 Displacement (mm) 0.23 mm (c) 20 stiffness (w/o spring) stiffness (w/ spring) loss factor (w/o spring) loss factor (w/ spring) (d) 20 0.2 0.2 15 0.1 N/mm 10 η (tanδ) 15 N/mm 0.4 10 0.1 η (tanδ) Force (N) 15 Fs1 Fs0 0.2 η (tanδ) (a) 8 5 5 0 0 –60 0 0 5 10 frequency (Hz) stiffness (w/o spring) loss factor (w/o spring) 15 20 0 –40 –20 0 20 40 60 80 temperature (°C) stiffness (w/ spring) loss factor (w/ spring) stiffness loss factor Fig. 14 (a) Displacement-load plot of single isolator strut. (b) Displacement dependency (15 Hz, 20°C). (c) Frequency dependency (0.23 mm, 20°C). (d) Temperature dependency (without spring, 15 Hz, 0.23 mm). is shown in Fig. 14(b). The test was conducted with two configurations: one with the coil springs and one without them. As the displacement increases, the stiffness decreases, and the loss factor increases. As the coil springs have no damping property, the strut with the coil springs has a lower loss factor than the strut without the coil springs. The frequency dependency, shown in Fig. 14(c), is relatively small, within the 5- to 20-Hz range. The temperature dependency, obtained only for the strut without coil springs, is shown in Fig. 14(d). As the temperature increases, the stiffness increases, and the loss factor decreases. At temperatures below −30°C, the SCSMA is in the temperature-induced martensitic state. The isolator strut exhibits damping properties at low temperatures, such as −54°C. The availability of damping materials that can be used at the same temperatures as this strut is limited. These data are used to correlate the structure math model, confirming that the VIS design should satisfy the transmissibility requirements shown in Fig. 4. J. Astron. Telesc. Instrum. Syst. 042004-11 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM An accelerated lifetime test was conducted on the SCSMA leaf spring. The spring was subjected to deformation at 200 Hz, which is 13.3 times the expected frequency in orbit. After three months of testing, no significant changes were observed in its properties, indicating a lifetime of more than 3 years in orbit. This confirms that the requirements stated in Sec. 2.1.1 have been met. 3.3 Hold and Release Mechanism The HRM holds the CMP securely during launch to prevent excessive movement and avoid VIS malfunction. The HRM for the XRISM VIS is integrated into the DWR plate and is in either the locked state or the released state, as depicted in Fig. 15. In the locked state [Fig. 15(b)], the CMP and DWR plates are rigidly connected, allowing the VIS to withstand high acceleration. The inner conical fitting attached to the DWR plate presses against the outer conical fitting at the four corners of the CMP plate, securing it in place. The preload, which pushes the inner conical fitting against the outer conical fitting, is a critical parameter in the design of the HRM. The VIS design should ensure the following: • During launch or ground tests, the preload must exceed the maximum force applied to the engaged fittings. Otherwise, a gap may form between the outer and inner conical fittings during vibration, which increases CMP acceleration. • The structure of the HRM, including the CMP plate and the DWR plate, should be stiff enough to prevent gaps in the fittings under the applied load and strong enough to avoid permanent deformation. • The HRM reset mechanism should generate sufficient force to apply the preload. In addition, consideration should be given to the material used in the HRM, the size of the motor and pin-puller, and the overall mechanism design to meet these conditions. In the released state [Fig. 15(c)], three isolator bipods suspend the CMP from the DWR plate, reducing vibration transmission from the CMP to the DMS. The transition from the locked state to the released state is triggered by activating the EBAD TiNi™ auto-resettable P5-5004 pin-puller, which can be remotely reset. When the pin-puller is activated, the drive shaft moves to the released position due to spring-restoring force, separating the outer and inner conical fittings, and putting the VIS in the released state. Conversely, the HRM reset mechanism returns the VIS to the locked state. Its motor moves the drive shaft from the released to the locked position. Subsequently, the pin-puller operates to (a) Auto-resettable pin-puller 2ST compressor (CMP) Geared motor CMP plate CMP fixes to the CMP plate DWR plate is fixed to the DWR Isolator bipod CMP is supported by 3 bipods in released state (b) CMP side Conical-conical fitting DWR side Motor arm Drive shaft (c) Pin-puller rod Compression spring Fig. 15 HRM: (a) Schematic view. (b) Locked state. (c) Released state.18 J. Astron. Telesc. Instrum. Syst. 042004-12 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM lock the mechanism, and the pin-puller rod fits into the drive shaft groove, securing the VIS. Finally, the motor arm moves to the right side to avoid interference when the VIS is released. GSE controls the HRM reset mechanism, as there is no need for in-orbit operation. Note that the simple release operation by a pin-puller and springs provides robust release in orbit, in contrast to the more sophisticated on-ground reset operation utilizing a motor and gears. The HRM reset mechanism is designed to lock and release multiple times on the ground, facilitating Resolve’s performance evaluation. The HRM was designed so that the mechanism can be held in place by the motor alone, without resetting the pin-puller, whereas the pin-puller can be locked as necessary. Both operations can be performed electrically without human access. In addition, two heaters (primary and redundant) were attached to a pin-puller for temperature control as the requirement in Sec. 2.2 specifies operation in the range of −30°C to þ 40°C. 3.4 Thermal Strap 3.4.1 Function and design of TS The TS is a heat path used to dissipate the heat generated by the CMP. It is designed to be flexible to prevent the transmission of CMP vibrations and is composed of multi-layer graphite sheets (GS) and aluminum end fittings. The GS material used is called Graphinity™, which is manufactured by Kaneka Corporation. Graphinity™ is known for its lightweight (with a density of 1.8 g∕cm3 ) and high thermal conductivity (three times that of copper). In addition, it is soft, easy to bend, and has low outgassing properties. Three types of TSs, namely A, B, and C, have been developed for the XRISM VIS, as shown in Fig. 16. Thermal conductivity and stiffness tests were performed on these types. Type A was used in conjunction with PC-B, which connects the CMP and the satellite base’s cold plate. Types B and C, on the other hand, were used with SC-A, SC-B, and PC-A, which connect the CMP and the DMS. 3.4.2 Verification of TS The thermal conductance of the TSs was measured in the configuration shown in Fig. 17. One end of the TS, which is expected to be hotter in orbit, was attached to a heater, whereas the other Fig. 16 Three types of TS: (a) Type A. (b) Type B. (c) Type C. Heater TS Cold plate Vacuum chamber (a) (b) Fig. 17 Thermal conductivity test configuration of TS: (a) Measurement setup. (b) Vacuum chamber. J. Astron. Telesc. Instrum. Syst. 042004-13 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 4 Results of thermal conductance tests of individual TS. ΔT (K) Thermal conductance (W/K) Requirement (W/K) Assigned to Type Phase No. Heater power (W) A FM 1 29.7 21.6 1.37 1.00 PC-B B QM 1 29.7 35.8 0.83 0.75 — QM 2 29.7 30.1 0.99 — FM 1 29.7 33.7 0.88 SC-A FM 2 29.7 37.3 0.80 SC-B FM 3 29.7 30.9 0.96 PC-A FM 4 29.7 32.4 0.92 SC-B FM 5 29.7 30.4 0.98 PC-A FM 6 29.7 32.9 0.90 SC-A QM 1 22.4 32.0 0.70 QM 2 22.4 36.4 0.61 — FM 1 22.4 34.4 0.65 SC-B FM 2 22.4 33.0 0.68 SC-B FM 3 22.4 32.3 0.69 PC-A FM 4 22.4 31.6 0.71 PC-A FM 5 22.4 34.7 0.64 SC-A FM 6 22.4 35.0 0.64 SC-A C — 0.62 end was attached to a cold plate. The measurement was conducted in a vacuum chamber, and the thermal conductance was calculated as the ratio of the heater power to the temperature difference between both ends. All TSs were found to meet the requirement stated in Sec. 2.1.3. Table 4 shows the results of the thermal conductance tests of TSs. As PC-A was the most thematically critical, the most conductive straps were assigned to it. The remaining straps were assigned to SC-A and SC-B to ensure that each system’s thermal conductance became nearly equal. The stiffness of the TSs was measured using the configuration shown in Fig. 18. A vibrator was attached to one end of the TS, and the displacement was measured with a laser displacement sensor. The reaction force at the opposite end of the TS was also recorded. The stiffness of the TSs was determined as the ratio of the measured force to the displacement. It was confirmed that the stiffness of the TSs was significantly lower than that of the isolator struts, ensuring that the Laser displacement sensor (a) Force sensor (b) TS Vibrator (c) Fig. 18 Stiffness test configuration of TS: (a) Measurement in the longitudinal direction. (b) Measurement in the traverse direction. (c) Measurement at low temperature (−30°C). J. Astron. Telesc. Instrum. Syst. 042004-14 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM vibration isolation performance was not impeded. The stiffness measurement test for the BBM was conducted at low temperature. The TS was cooled using a freeze spray, as shown in Fig. 18(c), and the stiffness was measured when it reached −30°C. The stiffness was found to be 1.3 times higher at −30°C compared with 20°C (room temperature). However, this increase is not considered significant in terms of vibration isolation performance. In addition, stiffness was found to exhibit minimal frequency dependence. 4 Verification of VIS 4.1 Overview During the manufacturing phase, five VISs were produced, including one qualification model (QM) and four flight models (FMs): SC-A, SC-B, PC-A, and PC-B. The following verification tests were conducted for all models. Although vibration transmissibility, thermal conductance, and mechanical endurance are not significantly affected by gravity, the displacement of the released HRM is greatly influenced by it. Therefore, the HRM release operation under 1G was verified through the HRM release operation test, and the operation under 0-G was verified by a combination of this test and analysis. 4.2 Vibration Transmissibility 4.2.1 Method of measurement The requirement for vibration transmissibility is defined as the ratio of the input force/moment at the CMP to the reaction force/moment of the fixed end, as described in Sec. 2.1.2. However, fixing one end of the VIS in inertial space is quite challenging, especially when measurements at various temperatures and directions of gravity is necessary. Therefore, the transmissibility was calculated as the ratio of the base acceleration/angular acceleration to the CMP acceleration/ angular acceleration. The base fixture was excited using an inertial vibrator, and the acceleration of a CMP simulator was measured, as shown in Fig. 19. Two inertial vibrators were used for each measurement, as indicated by the blue arrows; the red arrows show the movements of the CMP. For the translation measurement, two vibrators were driven in the same direction, whereas for the rotation measurement, they were driven in opposite directions. The following explanation outlines how the same transmissibility can be obtained by measuring the ratio of acceleration to acceleration instead of the ratio of force to force. Let us consider a single-degree-of-freedom (1-DOF) vibration isolator model, as depicted in Fig. 20. Figure 20(a) shows a fixed base model where the excitation force acting on the mass is a sinusoidal force (Fmi expðjωtÞÞ with angular frequency ω. The equation of motion and the reaction force at the base (Fmo expðjωtÞ) are as follows: X translation Y translation Z translation Y rotation Z rotation Z X Y VIS local coordinate X rotation Fig. 19 Transmissibility test configurations. J. Astron. Telesc. Instrum. Syst. 042004-15 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM (a) exp exp (b) mm km exp mm cm cm km exp exp Fig. 20 1-DOF model of vibration isolator: (a) fixed base configuration and (b) base excitation configuration. ð−mm ω2 þ jcm ω þ km ÞX a expðjωtÞ ¼ Fmi expðjωtÞ; (3) Fmo expðjωtÞ ¼ ðjcm ω þ km ÞX a expðjωtÞ: (4) EQ-TARGET;temp:intralink-;e003;114;594 EQ-TARGET;temp:intralink-;e004;114;564 Here, mm , cm , km , and X a represent mass, damping coefficient, spring constant, and displacement amplitude of the mass, respectively. The vibration transmissibility τm, which is the ratio Fmo to Fmi , becomes as follows: τm ¼ EQ-TARGET;temp:intralink-;e005;114;521 Fmo jcm ω þ km ¼ : Fmi −mm ω2 þ jcm ω þ km (5) On the other hand, for the base excitation model shown in Fig. 20(b), the equation of motion is given by −mm ω2 Xb ejωt þ jcm ωðX b − Y b Þejωt þ km ðXb − Y b Þejωt ¼ 0: EQ-TARGET;temp:intralink-;e006;114;460 (6) Here, X b and Y b represent the displacement amplitude of the mass and the base, respectively. From Eqs. (5) and (6), the ratio of the mass acceleration (ω2 X b ) to the base acceleration (ω2 Y b ) is as follows: EQ-TARGET;temp:intralink-;e007;114;399 ω2 X b X b jcm ω þ km ¼ ¼ ¼ τm : ω2 Y b Y b −mm ω2 þ jcm ω þ km (7) As explained above, the two test methods are mathematically equivalent. However, there are a few points that should be mentioned: • The base fixture should be sufficiently rigid to prevent resonance within the measurement frequency range. • In general, vibrators and accelerometers perform poorly at low frequencies. It should be noted that measurements at low frequencies may be challenging. • The VIS might have displacement dependency. Therefore, to compare vibration transmissibility measured by different methods, it should be measured at the same relative displacement in the same direction. 4.2.2 Measurement result Vibration transmissibility tests were conducted on five VISs at three temperatures: room temperature (RT), low temperature (−30°C), and high temperature ( þ 40°C). Due to the characteristics of the vibrator and the accelerometers, the measurement frequency range was limited to 25 to 600 Hz. The base fixture was designed to be sufficiently rigid to avoid interfering with the test results. The vibration transmissibility requirement must be met in orbit and under 1 G on the ground. However, the tests were conducted exclusively under 1 G conditions because testing under 0 G conditions is challenging in a ground-based environment, and the vibration transmissibility is not to be affected by gravity as long as the springs used in the vibration isolator have linear characteristics. Specifically, the SC-A and SC-B VIS units are mounted on the side of the DMS, with gravity acting in the Y direction of the local coordinate system. Conversely, the PC-A and PC-B units are mounted on the bottom of the DMS, experiencing gravity in the −Z direction. These J. Astron. Telesc. Instrum. Syst. 042004-16 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Fig. 21 Transmissibility test configurations: (a) SC-A and SC-B. (b) PC-A. (c) PC-B. Fig. 22 Vibration transmissibility of the qualification model (QM). configurations are shown in Fig. 21. For the QM, the test was performed with gravity acting in the þ Z direction (not shown in Fig. 21). Notably, the transmissibility did not exhibit significant differences based on the direction of gravity. Figure 22 illustrates the transmissibility of the QM across all temperatures, confirming that the requirement stated in Sec. 2.1.2 was met. All FMs also met the requirement. 4.3 Thermal vacuum test The thermal conductance of each VIS was verified with a full set of TSs. The test was conducted in a vacuum chamber under the expected heat load in orbit. Heat generation from the CMP was simulated using a heater attached to the CMP simulator. All VISs met the requirement specified in Sec. 2.1.3, as shown in Table 5. Thermal cycle tests were conducted in a vacuum chamber for all the FMs and the QM. First, the VISs were exposed to the lowest and highest storage temperatures (−60 to þ 40°C). Second, an eight-round thermal cycle test was conducted between the lowest and highest operating temperatures (−30 to þ 40°C). After these tests, all the VISs operated normally and met the requirements stated in Sec. 2.2. 4.4 HRM Function The operation of the pin-pullers’ release was verified in a vacuum chamber. The tests were conducted under worst-case conditions, including a temperature of −30°C and a current of 1.0 A. The pin-puller operates using the shape memory effect of the SMA, which is triggered by a rise in temperature when an electrical current is applied. Therefore, operation at the minimum temperature and current is considered the most challenging. All of the pin-pullers met the requirements in Sec. 2.3, as shown in Table 6. J. Astron. Telesc. Instrum. Syst. 042004-17 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 5 Results of thermal conductance test. Item Requirement (>W/K) Thermal conductance (W/K) Heat load (W) High temp. (°C) Low temp. (°C) SC-A 2.0 2.42 63.2 40.3 14.2 2.44 34.9 39.5 25.2 2.46 63.6 42.9 17.0 2.49 34.6 40.6 26.7 2.73 63.3 37.2 14.0 2.78 35.0 37.7 25.1 1.04 64.4 37.0 -25.1 1.05 34.9 38.6 5.3 2.14 63.7 41.4 11.6 2.14 34.6 42.0 25.9 SC-B PC-A PC-B QM 2.0 2.0 1.0 2.0 Table 6 Results of the pin-puller operation test under a temperature of −30°C. Current (A) Duration (ms) Item Result Requirement Result Requirement Result SC-A 1.0 1.0 to 1.6 14.0 <29 Pass SC-B 1.0 18.4 Pass PC-A 1.0 16.4 Pass PC-B 1.0 17.2 Pass QM 1.0 19.2 Pass displacement (mm) The displacements of the CMP were verified. The displacements under 1G were measured using laser displacement sensors. When the lock was released, the CMP was pulled by gravity, causing it to oscillate with damped motion. An example of the displacement time history is shown in Fig. 23, demonstrating that the movement of the CMP stabilizes in ∼1 s. The displacements under 0G were verified based on the 1G test and the 1G static analysis of the VIS. The static analysis indicates a displacement of 0 mm at 0G, whereas the displacement of the FMs at 0G is adjusted to minimize the displacement at 1G. Therefore, it can be estimated that the difference between the 1G static analysis and the settled displacement at 1 G represents the 2 1.6 1.2 0.8 0.4 0 0 1 2 3 time (s) 4 5 6 Fig. 23 SC-A CMP displacement in the Y direction at the capillary mounting point. J. Astron. Telesc. Instrum. Syst. 042004-18 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 7 CMP displacement during the releasing operation. Item Measured point SC-A Capillary attachment point LHP attachment point SC-B Capillary attachment point LHP attachment point PC-A PC-B Capillary attachment point Capillary attachment point Analysis 1G Disp. 1G Disp. 0 G Disp. (1G static) Requirement (max) (steady) (estimated) Direction (mm) (mm) (mm) (mm) (mm) Y 2.19 ±7.8 1.81 1.12 −1.07 Z 0.38 ±6.7 0.41 0.25 −0.13 Y 2.74 ±5.0 2.27 0.73 −2.01 Z 0.61 ±1.0 0.66 0.41 −0.20 Y 2.19 ±7.8 1.76 1.11 −1.08 Z 0.38 ±6.7 0.45 0.27 −0.11 Y 2.74 ±5.0 2.20 0.79 −1.95 Z 0.61 ±1.0 0.73 0.44 −0.17 Y −0.07 ±1.9 −1.07 −0.64 0.43 Z 0.77 ±3.4 1.15 0.72 0.05 Y −1.07 ±3.4 −0.92 −0.31 0.76 Z 0.77 ±8.3 1.62 1.07 0.30 displacement of the FM at 0G. All results are summarized in Table 7, where “1G Disp. (max)” refers to the peak displacement, “1G Disp. (steady)” indicates the settled displacement, and “0G Disp. (estimated)” represents the estimated displacement under 0G. As shown, all displacements comply with the requirements outlined in Sec. 2.1.5. For QM, a lifetime test of the HRM mechanism was conducted. As described in Sec. 2.4, the HRM must be capable of being locked and released at least 20 times during ground testing without disassembling the satellite. During QM testing, after the locking and release process was performed at least 40 times (twice the requirement), it was confirmed that the mechanism was still in good condition. 4.5 Vibration Test Vibration tests were conducted to verify the mechanical endurance of the VIS during launch, especially when the HRM is locked. These tests also confirmed that the acceleration at the CMP did not exceed its acceptable level during the AT. The tests were performed using a CMP simulator that emulated the mass properties of the actual CMP. Both sine and random vibration tests were conducted in the X, Y , and Z directions. Modal tests were carried out before and after each vibration test to ensure that no damage occurred. Figure 24 shows the frequency response function (FRF) of the QM VIS during the modal test. The FRF, which is the ratio of the acceleration measured at the CMP simulator to the base, shows that there was no significant amplification under 100 Hz. Table 8 and Fig. 25 show the results of Fig. 24 Frequency response function of VIS QM model at the CMP simulator. J. Astron. Telesc. Instrum. Syst. 042004-19 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 8 Modal analysis in a locked state. Effective mass ratio (%) Mode Hz X Y Z Rx Ry Rz 1 111.7 0 26 0 91.5 0 6.2 2 141.0 40.1 0 0 0 63.3 0 3 159.3 0 0 81.1 0 10.1 0 4 195.5 39.8 0 0 0 0.3 0 5 254.5 0 54.8 0 1.3 0 13 6 474.1 0 0 0 0 2.2 0 7 586.6 0 0 0 0 0 20.6 8 587.6 0 0 4.5 0 5.1 0.1 9 738.2 0.5 0 1.4 0 2.5 0 10 742.0 0.1 0 1.2 0 0.1 0 11 766.2 0.4 0 2.7 0 0.1 0 12 893.9 0 0 1.2 0 2.7 0 81.5 90.1 95 95.8 88 50 Total (%) (a) (b) Mode 1 111.7 Hz (c) Mode 2 141.0 Hz Mode 3 159.3 Hz Fig. 25 Modal shapes in launch configuration. (a) Mode 1 111.7 Hz. (b) Mode 2 141.0 Hz. (c) Mode 3 159.3 Hz. the modal analysis in the locked state and the modal shapes of the first three modes, respectively. The CMP was modeled as a single mass. The peaks in Fig. 24 are likely to correspond to these modes. Sine vibration tests were performed on the FMs at the AT level, whereas the QM was subjected to vibration at the QT level, which is 1.25 times as high as the AT level. The random vibration test level was the same for both QM and FM, but the duration was twice as long for QM than for FM. For the QM, one article was tested at a time (Fig. 26). However, to expedite the test campaign, two FMs were tested simultaneously (Fig. 27). All the articles met the requirements outlined in Secs. 2.1.4 and 2.2. 4.6 Verification and Validation at the Instrument and Spacecraft Levels The four sets of VIS were integrated into the DWR, and verification was conducted at the instrument level. Following the integration of the Resolve instrument, a series of spacecraft-level verification tests, including electrical, thermal, and mechanical tests, were carried out. These verifications at both the instrument and spacecraft levels confirmed the performance and J. Astron. Telesc. Instrum. Syst. 042004-20 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Fig. 26 Vibration test configuration for QM VIS. Fig. 27 Vibration test configuration for FM VIS. functionality of the VIS. The system’s vibration isolation efficacy was substantiated by measurements of the vibration levels on the DMS and analysis of the detector noise spectra and energy histograms derived from calibration with X-ray sources. In addition, the release and locking functions of the HRM were tested repeatedly. The thermal performance of the TS was demonstrated during the spacecraft thermal-vac test. 4.6.1 Vibration level on the DMS To verify the effect of the VIS before installing the detector in the DWR, the allowable vibration level at the representative point of the DMS was defined as a requirement based on the Hitomi results.18 This requirement applies not only to the VIS but also to the entire DWR system, which includes the CMPs, CHs, and VISs. The allowable vibration level in the Z direction (DWR local coordinate system shown in Fig. 1) at the top of the middle of the DMS (between SC-A and SCpffiffiffiffiffiffi B) was defined as 2 × 10−4 G∕ Hz or lower within a frequency range of 100 to 600 Hz. This requirement was applied to the continuum component and not to the line components at the harmonics of the cryocooler’s drive frequency. As shown in Fig. 28, the continuum component pffiffiffiffiffiffi of the vibration level when the cryocoolers were operated was below 1 × 10−4 G∕ Hz, which satisfies the requirement. Both CMPs and CHs produce many harmonics of their driving frequency of 15 Hz. However, the harmonics shown in Fig. 28 are mostly caused by CHs. Although CMPs are mounted on the VIS, CHs are rigidly mounted on the DWR. Note that these harmonics do not degrade the detector performance as long as the frequencies are off the natural frequency of the CSI or the detector. 4.6.2 Detector performance The four VIS units were installed on the DWR before the detector performance test. The performance of the VIS was demonstrated by measuring the detector noise and energy resolution. In all the detector performance tests on the ground, regardless of ambient or thermal-vac environment (i.e., during the spacecraft thermal-vac test), the noise feature was free from vibration disturbance when the VIS was released, validating the performance of the VIS. The energy resolution of the detector was <5 eV at 5.9 keV, satisfying the ground test allocation (<5.9 eV) for the in-orbit requirement of <7 eV. The detector noise and energy resolution J. Astron. Telesc. Instrum. Syst. 042004-21 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Fig. 28 Power spectral density (PSD) of the acceleration measured on the DMS when all the coolers are operating.18 performance were also measured with the VIS in the locked state, where vibration isolation was not applicable. The noise was excessive at low frequencies, and the energy resolution was degraded. Examples of the noise spectra and energy histogram of the Fe-55 calibration source detected by detector pixels are shown in Fig. 29. Figures 29(a) and 29(b) show the data taken with the VIS released, whereas Figs. 29(c) and 29(d) show those with the VIS locked. The noise spectra represent the Fourier-transformed detector voltage signal when no X-ray hits occur. When the VIS was locked and the cryocooler vibration was not isolated, the noise spectrum exhibited excess in the low-frequency band, affecting the detector’s energy resolution. The low-energy excess suggests that the detector and/or its wiring stochastically move, leading to variations in detector gain. The two peaks of the energy histogram in panel (b) correspond to the Ka1 and Ka2 transitions of Mn Fig. 29 Comparison of noise spectra (a), (c) and energy histogram (b), (d). Panels (a) and (b) are taken with the VIS released, whereas panels (c) and (d) are taken with the VIS locked. Blue, red, green, and magenta lines in panels (a) and (c) show the noise spectrum of pixels 0, 1, 2, and 3, respectively. Energy histograms (b), (d) are taken with the Fe-55 calibration source. Data from all detector pixels are combined. J. Astron. Telesc. Instrum. Syst. 042004-22 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 9 Number of operations of the HRM. Environment On-ground (instrument-level and spacecraft-level) On-orbit HRM state SC-A SC-B PC-A PC-B Pin-puller release 23 24 30 26 Lock (motor only) 7 7 8 7 Lock (pin-puller) 23 24 30 26 Pin-puller release 1 1 1 1 Lock (motor only) N/A N/A N/A N/A Lock (pin-puller) N/A N/A N/A N/A atoms produced by electron capture of the Fe-55 radioisotope. The distinction between the two peaks is unclear in panel (d) due to the degradation of energy resolution from ∼4.0 to ∼6.5 eV. 4.6.3 HRM function The HRM was repeatedly released and locked. When testing the detector’s performance, the HRM was in the released state. By contrast, the HRM had to be locked during mechanical tests of the DWR and spacecraft. In addition, as the CMP shifts ∼2 mm under 1 G, the HRM should be locked to prevent excessive movement when relocating the DWR or the spacecraft. Once the DWR was integrated into the spacecraft, the VIS became inaccessible to the operational staff. The HRM was designed to be electrically reset to the locked position, obviating the need for manual intervention. Table 9 summarizes the HRM operations, underscoring the significance of the reset function without needing to replace the pin-puller unit. Without this function, the financial and scheduling repercussions would be considerable, possibly impeding the comprehensive validation of detector performance. The HRM can secure the CMP solely with its motor mechanism, bypassing the need to reset the pin-puller. The pin-puller was reset for configurations akin to actual flight conditions (e.g., mechanical tests, thermal-vac tests, and HRM functionality tests), whereas the motor-lock setup was employed to secure the mechanism during spacecraft transport operations. The reset process was intended to be executed using specialized GSE. The XRISM VIS cannot reset the HRM while in orbit. The spacecraft’s thermal-vac test offered the optimal opportunity to demonstrate the HRM’s release feature in an in-orbit-like environment. As the pin-puller employs an SMA, its energy margin decreases at lower temperatures. As a demonstration in the worst case, the release procedure was conducted at a low spacecraft temperature, with the heater regulated to align with inorbit operation protocols. The successful release of the four HRM units validated their functionality and their compatibility with the spacecraft’s drive electronics. 4.6.4 Thermal strap performance The thermal performance of the VIS, achieved by the TS units, was demonstrated during the spacecraft thermal-vac test. However, due to the limited number of temperature sensors, accurate measurements of each TS were not possible. Table 10 shows the temperatures of the four CMPs, the allowable temperature ranges, and the expected values from the math-model-based analysis. The data shown correspond to the hot case with nominal cryocooler power. The CMP temperatures were within the acceptable range with a fair margin, thereby demonstrating the performance of the TSs. 4.6.5 On-orbit performance The HRM of the VIS was successfully released on September 7, 2023, at 0:33 UT, ∼50 min after launch, with the temperature of the pin-pullers controlled at 30°C. The Resolve detector was J. Astron. Telesc. Instrum. Syst. 042004-23 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Table 10 Temperature measured during the spacecraft thermal-vac test (hot case). Position Measured temperature (°C) Analysis prediction (°C) Allowable range (°C) SC-A CMP +0.4 −3.0 −40/+40 SC-B CMP +1.6 −3.0 −40/+40 PC-A CMP +27.8 +23.0 −40/+40 PC-B CMP +14.7 +9.0 −40/+40 Fig. 30 (a) An in-orbit noise spectrum of the Resolve detector. (b) Energy histogram.29 cooled to its operating temperature of 50 mK on October 9, 2023. The detector’s performance was then verified during the commissioning period of the mission. The detector’s energy resolution met the requirement of 7 eV, achieving ∼5 eV. The temperatures of the CMPs are all within the allowable range by a fair margin. The typical temperatures of each CMP measured in orbit are −10°C for SC-A and SC-B, þ 10°C for PC-B, and þ 25°C for PC-A, with a variation of ± a few °C depending on the orbital motion and pointing direction. The temperature of the VIS should be similar to that of the associated CMP. The VIS functions in all required aspects. Figure 30 shows an example of the noise spectra and the detector energy histogram taken in orbit. 5 Lessons Learned The following lessons were learned through the development of the XRISM VIS and are helpful when developing a VIS to decouple microvibrations from cryocoolers with a sensitive detector. • In general, highly sensitive sensors are also susceptible to noise sources such as mechanical vibrations, a broad band of electromagnetic waves, and others. The most reliable way to determine noise immunity is to test using real (similar) hardware. Therefore, we strongly recommend developing an engineering model (EM) of the sensor in the early stage of development. In the case of Hitomi, the mechanical noise issues were not recognized until testing was conducted using the EM. For XRISM, as the designs of the CSI and the DWR were quite similar, the same threshold level could be assumed. • In the case of Hitomi, the decision to adopt the VIS was made during the manufacturing phase. As a result, the interface between the VIS and the DWR had to be aligned with the already manufactured DWR, imposing rigid constraints on the design of the VIS. In contrast, for Resolve, although the fundamental DWR design is the same as that for the SXS, the decision to adopt VIS was made before the manufacturing of DWR, allowing for the redesign of the DWR to create an interface tailored to the VIS. This significantly increased the design flexibility of VIS and provided a slight buffer in the development schedule. If the DWR could have been newly designed, the VIS would have been designed much more flexibly. If mechanical vibrations could cause problems, the need for VIS should be J. Astron. Telesc. Instrum. Syst. 042004-24 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM determined early in the project. If uncertainty remains, the development should proceed with the incorporation of the VIS as it can be easily removed afterward. • The detector’s performance in the in-orbit configuration can only be measured when the HRM (i.e., the launch-lock mechanism) is released. However, it must be locked for mechanical verification or other purposes (e.g., for protection from damage due to large movement under 1 G). As these two requirements conflict, the capability to re-lock the mechanism without replacing any parts is necessary. The design described in this paper provides one such solution. • As the pin-puller is based on an SMA, the energy required to release it depends strongly on the temperature of the pin-puller housing. To reduce uncertainty and secure the energy margin, the thermal design of pin-pullers, which includes thermal isolation and heater control, is crucial. • An HRM reset failure occurred during one of the spacecraft-level tests. Although it was impossible to directly observe the VIS, the cause of the failure was determined to be a slight positional deviation of the HRM’s parts as identified through a fault tree analysis. The following measures should be taken to avoid such failures: (1) The tolerances for moving parts should be large enough to accommodate slight deviations in their relative positions. (2) For components that bear a large load in the shear direction of the fastening bolts, positioning should be done using shear pins in conjunction with the bolts. 6 Conclusion The design and performance of the VIS for the Resolve instrument onboard the XRISM spacecraft are described. The VIS was designed to mitigate the degradation of the spectroscopic performance of a cryogenic microcalorimeter instrument due to mechanical disturbances from cryocooler compressors. The VIS includes a vibration isolation function, a heat transfer function, and hold-and-release functions. These functions work together to keep the cryocoolers within their allowable environment and ensure that there is no negative impact on the detector’s performance. We hope that the design solutions and lessons described in this paper can serve as a valuable reference for future cryogenic space missions, particularly those requiring unprecedented high sensitivity. Disclosures The authors declare the following financial interests and/or personal relationships that could be considered potential competing interests: Kazuki Watanabe holds patents related to a shockabsorbing device issued to WEL Research. All other authors declare that they have no known competing financial interests or personal relationships that could have influenced the work reported in this paper. Code and Data Availability The data supporting the findings of this article originate from internal reports generated during the project and are not publicly available. Code sharing is not applicable to this article as no new code was developed. Acknowledgments The authors express their gratitude to all members of the Hitomi SXS team and the XRISM Resolve team. During the preparation of this work, the authors utilized Microsoft Copilot and OpenAI ChatGPT to enhance the readability and language of the paper. After using these services, the authors thoroughly reviewed and edited the content as needed and take full responsibility for the published article’s content. References 1. N. W. Boggess et al., “The COBE mission: its design and performance two years after launch,” Astrophys. J. 397, 420 (1992). 2. C. L. Bennett et al., “The microwave anisotropy probe mission,” Astrophys. J. 583(1), 1 (2003). J. Astron. Telesc. Instrum. Syst. 042004-25 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM 3. J. A. Tauber et al., “Planck pre-launch status: the Planck mission,” Astron. Astrophys. 520, A1 (2010). 4. H. Murakami et al., “The infrared astronomical mission AKARI,” Publ. Astron. Soc. Jpn. 59, S369–S376 (2007). 5. M. W. Werner et al., “The Spitzer space telescope mission,” Astrophys. J. Suppl. Ser. 154(1), 1–9 (2004). 6. G. L. Pilbratt et al., “Herschel space observatory. An ESA facility for far-infrared and submillimetre astronomy,” Astron. Astrophys. 518(L1), 1–6 (2010). 7. J. P. Gardner et al., “The James Webb Space Telescope mission,” Publ. Astron. Soc. Pac. 135, 068001 (2023). 8. T. Takahashi et al., “Hitomi (ASTRO-H) X-ray astronomy satellite,” J. Astron. Telesc. Instrum. Syst. 4(2), 021402 (2018). 9. T. Ghigna et al., “The LiteBIRD mission to explore cosmic inflation,” Proc. SPIE 13092, 1309228 (2024). 10. P. Peille, V. Albouys, and D. Barret, “The X-ray integral field unit architecture following the Athena mission rescope,” Proc. SPIE 13093, 130930U (2024). 11. R. P. Kraft et al., “The Line Emission Mapper (LEM) Probe mission concept,” Proc. SPIE 13093, 1309327 (2024). 12. J. Glenn et al., “PRIMA: the probe far-infrared mission for astrophysics,” Proc. SPIE 13092, 130920J (2024). 13. E. Peretz, “The black hole explorer: astrophysics mission concept engineering study report,” Proc. SPIE 13092, 130922F (2024). 14. Y. Ishisaki et al., “Status of Resolve instrument onboard X-ray imaging and spectroscopy mission (XRISM),” Proc. SPIE 12181, 121811S (2022). 15. M. Tashiro et al., “Status of X-ray imaging and spectroscopy mission (XRISM),” Proc. SPIE 11444, 1144422 (2020). 16. R. Kelley et al., “The Astro-H high resolution soft X-ray spectrometer,” Proc. SPIE 9905, 99050V (2016). 17. Y. Ezoe et al., “Cooling system for the Resolve onboard XRISM,” Cryogenics 108, 103016 (2020). 18. S. Yoshida et al., “Performance test results of a helium Dewar for the Resolve instrument onboard the XRISM,” Cryogenics 139, 103831 (2024). 19. P. J. Shirron and M. O. Kimball, “On-orbit performance of the Resolve adiabatic demagnetization refrigerator on XRISM,” Proc. SPIE 13093, 1309360 (2024). 20. Y. Takei et al., “Vibration isolation system for cryocoolers of soft X-ray spectrometer on-board ASTRO-H (Hitomi),” J. Astron. Telesc. Instrum. Syst. 4, 011216 (2018). 21. Y. J. Kasai et al, “JEM/SMILES observation capability,” Proc. SPIE 7474, 74740S (2009). 22. K. Otsuka et al, “Improvement of micro-vibration of a two-stage Stirling cryocooler,” Cryogenics 111, 103133 (2020). 23. S. Yasuda and K. Ishimura, “Method of determining specification for transmissibility of vibration isolator for ASTRO-H soft X-ray spectrometer (SXS),” in 13th Eur. Conf. Spacecr. Struct., Mater. & Environ. Test., Braunschweig, Vol. 727, p. 29 (2014). 24. K. Watanabe, “Shock absorbing device,” Japan Patent Appl. PCT/JP2014/069286 (2014). 25. K. Watanabe, “Shock absorbing device,” Europe Patent Appl. EP 14898062.6 (2014). 26. K. Watanabe, “Shock absorbing device,” US Patent Appl. US 15/413148 (2017). 27. H. Uchida et al., “Microvibration isolation system with superelastic alloy for space applications,” in AIAA SCITECH 2022 Forum, no. 2022-1462 (2021). 28. V. Martynov et al., “Superelastic and shape memory single crystal Cu-Al-Ni: fabrication and applications,” in Proc. Int. Conf. Shape Mem. and Superelast. Technol., Pacific Grove, California, ASM International, pp. 427–438 (2006). 29. F. S. Porter et al., “In-flight performance of the XRISM/Resolve detector system,” Proc. SPIE 13093, 130931K (2024). Susumu Yasuda is a manager for structural engineering at the Japan Aerospace Exploration Agency (JAXA). He received his BS and MS degrees in mechanical engineering from the University of Tokyo in 1991 and 1993, respectively. From 1993 to 2006, he worked at Canon Inc., and joined JAXA in 2006, where he contributed to the development of vibration isolators for Hitomi/SXS and XRISM/Resolve. His primary research interest lies in mechanical vibration in spacecraft. Yoh Takei is an instrument engineer at the Japan Aerospace Exploration Agency (JAXA). He received his BS, MS, and PhD degrees in science from the University of Tokyo in 2001, 2003, and 2006, respectively. With over 15 years of experience, he has developed cryogenic X-ray spectrometers for the SXS instrument on Hitomi and the Resolve instrument on XRISM. In addition to instrument engineering, he has also worked as an astrophysicist and spacecraft systems engineer. J. Astron. Telesc. Instrum. Syst. 042004-26 Oct–Dec 2025 • Vol. 11(4) Yasuda et al.: Vibration isolation system for cryocoolers on the XRISM Atsushi Okamoto is a manager in the thermal control technology division at JAXA. He received his BS, MS, and PhD degrees in aerospace engineering from Kyushu University in 2001, 2003, and 2018, respectively. His work includes the development of thermal control systems for spacecraft and the management of research and development activities on thermal control devices. Keiichi Yanagase is a researcher in mechanical environmental testing and structural engineering at JAXA. He received his BS and MS degrees in materials engineering from Tohoku University in 2005 and 2007, respectively, and his PhD in aerospace engineering from Tottori University in 2021. Kazuki Watanabe is the president and CEO of Well Research Inc. and an inventor of the Vibration Isolation System (VIS). He earned his BS degree in aerospace engineering from Tokai University in 1982, his MS degree in mechanical engineering from Waseda University in 2002, and his PhD from Waseda University in 2005. Seiji Yoshida is a principal engineer at Sumitomo Heavy Industries, Ltd. He earned both his BS and MS degrees in mechanical engineering from Nagaoka University of Technology in 1996 and 1998, respectively. His primary focus is on pioneering the development of advanced cryogenic equipment for space applications. J. Astron. Telesc. Instrum. Syst. 042004-27 Oct–Dec 2025 • Vol. 11(4)
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )