ARCHITECT c4000 System Service and Support Manual Manual Revision Number 204733-126 Front Matter Content Control Number 204734-126 © 2009, 2015 Abbott Laboratories, Abbott Park, IL. All rights reserved. Revision Log Click to view Chapter 4 Removal and Replacement and Chapter 5 Verification Procedure Revision History All other Revision History is located in the table below: REVISION DATE SECTIONS REVISED/ADDED TSBs INCORPORATED ISAs INCORPORATED 204733-126 Apr/2015 c4000 Revision 204733-126 Change Listing N/A N/A 204733-125 Nov/2014 c4000 Revision 204733-125 Change Listing N/A N/A 204733-124 Oct/2014 c4000 Revision 204733-124 Change Listing N/A N/A 204733-123 June/2014 c4000 Revision 204733-123 Change Listing N/A N/A 204733-122 Apr/2014 c4000 Revision 204733-122 Change Listing N/A N/A 204733-121 Sept/2013 c4000 Revision 204733-121 Change Listing N/A N/A 204733-120 May/2013 c4000 Revision 204733-120 Change Listing N/A N/A 204733-119 Jan/2013 c4000 Revision 204733-119 Change Listing N/A N/A 204733-118 Aug/2012 c4000 Revision 204733-118 Change Listing N/A N/A 204733-117 July/2012 c4000 Revision 204733-117 Change Listing N/A N/A 204733-116 July/2012 c4000 Revision 204733-116 Change Listing N/A N/A 204733-115 JUN/2012 c4000 Revision 204733-115 Change Listing N/A N/A 204733-114 Mar/2012 c4000 Revision 204733-114 Change Listing N/A N/A 204733-113 July/2011 c4000 Revision 204733-113 Change Listing N/A N/A 204733-112 April/2011 c4000 Revision 204733-112 Change Listing N/A N/A 204733-111 Jan/2011 c4000 Revision 204733-111 Change Listing N/A N/A 204733-110 Oct/2010 c4000 Revision 204733-110 Change Listing N/A N/A 204733-109 Sept/2010 c4000 Revision 204733-109 Change Listing N/A 128-002A 204733-108 May/2010 c4000 Revision 204733-108 Change Listing N/A 128-002A 204733-107 Jan/2010 c4000 Revision 204733-107 Change Listing N/A N/A 204733-106 Nov/2009 c4000 Revision 204733-106 Change Listing N/A N/A 204733-105 Oct/2009 c4000 Revision 204733-105 Change Listing N/A N/A 204733-104 Sept/2009 c4000 Revision 204733-104 Change Listing N/A N/A 204733-103 Aug/2009 c4000 Revision 204733-103 Change Listing N/A N/A 204733-102 Jul/2009 c4000 Revision 204733-102 Change Listing N/A N/A 204733-101 Apr/2009 ALL (NEW PUBLICATION) N/A N/A ARCHITECT c4000 Service and Support Manual (Version 204733-126) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Proprietary Information (Document Control Number 204735-103) The information, documents and related graphics published herein (the “Information”) are the sole property of Abbott Laboratories. Permission to use the Information is granted, provided: • The copyright notice appears on all copies • Use of the Information is for operation of ABBOTT products by Abbott trained personnel or informational use only • The Information is not modified in any way; and • No graphics are used separate from accompanying text. Each person assumes full responsibility and all risks arising from use of the Information. The Information is presented “AS IS” and may include technical inaccuracies or typographical errors. Abbott Laboratories reserves the right to make additions, deletions, or modifications to the Information at any time without any prior notification. Qualifications: • All samples (printouts, graphics, displays, screens, etc.) are for information and illustration purposes only and shall not be used for clinical or maintenance evaluations. Data shown in sample printouts and screens do not reflect actual patient names or test results. • The information was developed to be used by Abbott Laboratories trained personnel, by other persons knowledgeable or experienced with the operation and service of the product identified, under the supervision and with cooperation from Abbott Laboratories technical support or service representatives. • In no event shall Abbott Laboratories or its affiliates be liable for any damages or losses incurred in connection with or arising from the use of the information by persons not fully trained by Abbott Laboratories. This limitation shall not apply to those persons knowledgeable or experienced with the operation and service of the product identified, under the supervision and with cooperation from Abbott Laboratories technical sales and service representatives. • No part of this media may be reproduced, stored, retrieved, or transmitted in any form or by any means without the prior written permission of Abbott Laboratories. • No confidential relationship shall be established in the event that any user of the Information should make any oral, written or electronic response to Abbott Laboratories (such as feedback, questions, comments, suggestions, ideas, etc.). Such response and any information submitted therewith shall be considered non confidential, and Abbott shall be free to reproduce, publish or otherwise use such information for any purposes whatsoever including, without limitation, the research, development, manufacture, service, use, or sale of products incorporating such information. The sender of any information to Abbott is fully responsible for its content, including its truthfulness and accuracy and its non infringement of any other person's proprietary rights. • Abbott Laboratories is not engaged in rendering medical advice or services. • Updates to the Information may be provided in either paper or electronic format. Always refer to the latest documents for the most current information. All Abbott Laboratories product names and trademarks are owned by or licensed to Abbott Laboratories, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, trade dress, or product name may be made without the prior written authorization of Abbott Laboratories, except to identify the product or services of Abbott Laboratories. All other trademarks, brands, product names, and trade names are the property of their respective companies. All rights reserved. Except as permitted above, no license or right, express or implied, is granted to any person under any patent, trademark, or other proprietary right of Abbott Laboratories. ABBOTT LABORATORIES MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY KIND OR NATURE WITH RESPECT TO THE INFORMATION. ABBOTT LABORATORIES HEREBY DISCLAIMS ALL REPRESENTATIONS AND WARRANTIES, WHETHER EXPRESS OR IMPLIED, CREATED BY LAW, CONTRACT OR OTHERWISE, INCLUDING, WITHOUT LIMITATION, ANY WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, TITLE OR NON INFRINGEMENT. IN NO EVENT SHALL ABBOTT LABORATORIES BE LIABLE FOR ANY DAMAGES OF ANY KIND OR NATURE, INCLUDING, WITHOUT LIMITATION, DIRECT, INDIRECT, SPECIAL (INCLUDING LOSS OF PROFIT) CONSEQUENTIAL OR INCIDENTAL DAMAGES ARISING FROM OR IN CONNECTION WITH THE EXISTENCE OR USE OF THE INFORMATION, REGARDLESS OF WHETHER ABBOTT LABORATORIES HAS BEEN ADVISED AS TO THE POSSIBILITY OF SUCH DAMAGES. ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. ARCHITECT c4000 Service and Support Manual (Version 204733-107) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-102 Change Listing This page lists the changes from 204733-101 to Revision 204733-102. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-102. FRONT Title Change Front Matter Content Control Number to 204734-102. FRONT Rev Log Add 204733-102 change log. Revision Change Proprietary Information Control Number to 204735-102. 1 Gen. Data 1 2 Updated links to Biological Safety section. Troubleshooting 2 7 Changed Document Control Number 204737-102. Updated text in Card Cage section. Pre-Site Specification & Checklist 7 8 Changed Document Control Number 204736-102. Change Document Control Number 204741-102 Updated text in System Specification. Installation Change Document Control Number 204742-102 8 Updated ci4100 Installation Checklist. 8 Corrected paragraph format in c4000 Attach External Plumbing and Water Bath filter. 8 Corrected paragraph format in c4000 Module Upack-Installation A and B. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-103 Change Listing This page lists the changes from 204733-102 to Revision 204733-103. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-103. FRONT Title Change Front Matter Content Control Number to 204734-103. FRONT Rev Log Add 204733-103 Revision Log. 1 Gen. Data Change Document Control Number 204736-103 Revision 1 Update text in Mechanical Hazards 1 Update text in Spill Clean-up 1 Update text in Chemical Hazards, Sodim Azide 7 Pre-Site Specification & Checklist Change Document Control Number 204741-103 7 Updated text in System Specification, under Electrical, Power, and Heat Output Specification and in Module UPS Specification. 7 Updated text in Module UPS Specification. 7 Updated text in Pre-Site Checklist, under c4000 Module Electrical Supply Requirements. 7 Added Electrical Shock Hazard symbols to 60 Hz Power Measurements and 60 Hz Power Measurements 7 Added Chemical Hazard symbols to Liquid Waste c4000 (Options), Liquid Waste i1000SR and Pump Waste (if required) ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-104 Change Listing This page lists the changes from 204733-103 to Revision 204733-104. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-104. FRONT Title Change Front Matter Content Control Number to 204734-104. FRONT Rev Log Add 204733-104 Revision Log. Troubleshooting Change Document Control Number 204737-103 2 2 Revision Update MECH LED graphic. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-105 Change Listing This page lists the changes from 204733-104 to Revision 204733-105. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-105. FRONT Title Change Front Matter Content Control Number to 204734-105. FRONT Rev Log Add 204733-105 Revision Log. General Data Change Document Control Number 204736-104 1 Revision 1 Updated text in Mechanical Hazards 1 Updated graphic 7A_1074b in Cuvette Washing - Water & Wash Solution Aspiration and Drying, Vacuum Pumps 1 Replaced graphic 7A_1053a with 7A_1074b in Robotic Assembly Description 2 2 Troubleshooting Change Document Control Number 204737-104 Updated graphic Mixer Block Diagram to 7A_2043b ARCHITECT c4000 Service and Support Manual (Version 204733-105) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-106 Change Listing This page lists the changes from 204733-105 to Revision 204733-106. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-106 FRONT Title Change Front Matter Content Control Number to 204734-106 FRONT Rev Log Add 204733-106 Revision Log 8 Installation Change Document Control Number to 204742-103 Revision 8 Reconfigured entire Installation section 8 Added Standalone Installation Instructions 8 Added New c4 and New i1000SR Installation Instructions 8 Added New c4 and Existing i1000SR Installation Instructions 8 Added Existing c4 and New i1000SR Installation Instructions 8 Added Existing c4 and Existing i1000SR Installation Instructions ARCHITECT c4000 Service and Support Manual (Version 204733-106) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-107 Change Listing This page lists the changes from 204733-106 to Revision 204733-107. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-107 FRONT Title Change Front Matter Content Control Number to 204734-107 FRONT Proprietary Information Change Revision Number to 204735-102 FRONT Proprietary Information Updated trademark information 1 General Data Change Document Control Number to 204736-105 Revision Updated Biological Hazard area, Precaution with new text references about reporting incidents. Updated Sodium Azide information in Chemical Hazards Update Fluidics Diagram 7A_9002b.pdf Added Total Wiring Diagram 7A_9003b to Key Diagrams, M&Ds and Alignment/Adjustment Procedures table Updated Wiring Diagram 7A_9003b Updated Power Switches text and updated table text. Updated text in Mixer Assembly 2 Troubleshooting Change Document Control Number to 204737-105 Updated Mixer Driver Board table text and graphics 7A_2077a and 7A_2044b Updated voltage output in Main Power Supply Updated text in Power Supply Connections, Transformer table and AC table Updated text Power Supply Schematics Table Added Text to Power Transformer Cable Pinout Added text Circuit Protector to MC1 under Transformer Updated schematic 7A_9027b.pdf in Stirrer Motor and Sensor Schematic Diagram ARCHITECT c4000 Service and Support Manual (Version 204733-107) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-108 Change Listing This page lists the changes from 204733-107 to Revision 204733-108. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-108 FRONT Title Change Front Matter Content Control Number to 204734-108 General Data Change Document Control Number to 204736-106 1 Revision Replaced Key Diagrams, M&Ds and Alignment Procedure table. 2 Troubleshooting Changed Chapter 2 Troubleshooting Document Control Number to 204737-106 Replaced AC/DC Driver graphic 7A_2042a with 7A_2079a Added new link to Mixer Driver Board Functional Diagram table.7A_20778a Replaced Mixer Driver Board graphic 7A_2044a with 7A_2078a Corrected Cable Routing and Connector Pin-out TS table Text correction to RSH Functional Diagram Link table Added new link in RSH Key Diagrams, M&Ds and Alignment Procedures table. Added graphic 7A_9045a. 6 Planned Maintenance Changed Chapter 6 Planned Maint. Document Control Number to 204740-102 Added text to PM Procedure #1 table Added text to PM #1 Checklist 8 Installation Changed Chapter 8 Troubleshooting Document Control Number to 204742-104 Installation / Standalone Installation Checklist text correction Unpack-Remove Packing Tape, Remove Center Support Boards text correction Transformer Voltage-Install Power Cord Robotic Position Setup-Delete Lifting Hazard Caution, changed list order in Flush Lines (c4000) Check Wash Cup Flow-text correction Installation / New c4000 and New i1000SR Installation Checklist text correction ci4100 Integration-text correction, replaced graphic 7A_8304a with 7A_8305a SCC HW and Cabling Installation-added Electrical Shock Hazard Warning, replaced Connect Power for SCC Components text, replaced Install Printer text Transformer Voltage-Install Power Cord text correction Robotic Position Setup-Check Wash Cup Flow text correction Installation / New c4000 and Existing i1000SR Installation Checklist text correction Unpack- Remove Center Support Boards text correction Set i1000SR Jumper to Module 2-added text SCC HW and Cabling Installation-added Electrical Shock Hazard Warning, replaced Install Printer text, replaced Connect Power for SCC Components text Installation / Existing c4000 and New i1000SR Installation Checklist text correction c4000 Kits Package Checklist-text correct Carrier Plate ci4100 Integration-correct text for Install Barcode Reader and Bottle Spinner, Install Covers SCC HW and Cabling Installation-added Electrical Shock Hazard Warning, replaced Connect Power for SCC Components text Robotic Position Setup- Check Wash Cup Flow text correction Installation / Installation Checklist text correction Existing c4000 and Existing i1000SR c4000 Kits Package Checklist-text correct Carrier Plate ci4100 Integration-delete graphic 7A_8304a, correct text for Install Barcode Reader and Bottle Spinner SCC HW and Cabling Installation-added Electrical Shock Hazard Warning, replaced Connect Power for SCC Components text ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-109 Change Listing This page lists the changes from 204733-108 to Revision 204733-109. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-109 FRONT Title Change Front Matter Content Control Number to 204734-109 Installation Changed Chapter 8 Troubleshooting Document Control Number to 204742-105 Installation / Standalone Updated table to include added task to Install Nameplate and Install Decal. c4000 Install Accessories Added task to Install Nameplate and Install Decal and give c4000 Accessory Kit #2 information. Installation / New c4000 and New i1000SR Installation Checklist text correction ci4100 Integration-text correction, replaced graphic 7A_8304a with 7A_8305a SCC HW and Cabling Installation-added Electrical Shock Hazard Warning, replaced Connect Power for SCC Components text, replaced Install Printer text Transformer Voltage-Install Power Cord text correction Robotic Position Setup-Check Wash Cup Flow text correction 8 Revision ci4100 Integration Added task to Install Nameplate and Install Decal and give c4000 Accessory Kit #2 information. Installation / New c4000 and Existing i1000SR Updated table to include added task to Install Nameplate and Install Decal. ci4100 Integration Added task to Install Nameplate and Install Decal and give c4000 Accessory Kit #2 information. ARCHITECT c4000 Service and Support Manual (Version 204733-109) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-110 Change Listing This page lists the changes from 204733-109 to Revision 204733-110. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-110 FRONT Title Change Front Matter Content Control Number to 204734-110 General Data Changed Document Control Number to 204736-107 1 Revision Added new bullet statement plus a Caution and removed autoclave statement from Sodium Azide information in Chemical Hazards. 6 Planned Maintenance Document Control Number, change to 204740-103 Updated PM Procedure #1 table to “To check degasser output for bubbles”. Updated PM Procedure #2 table to “To check degasser output for bubbles”. Updated PM Procedure #3 table to “To check degasser output for bubbles”. Updated PM #1 Checklist table to add step “Check degasser output for bubbles” and renumber other steps. Updated PM #2 Checklist table to add step “Check degasser output for bubbles” and renumber other steps. Updated PM #3 Checklist table to add step “Check degasser output for bubbles” and renumber other steps. 7 Pre-Site Specification and Checklist Document Control Number, change to 204741-104 Added ‘Sodium Azide Awareness” statement to Pre-Site Checklist table. ARCHITECT c4000 Service and Support Manual (Version 204733-110) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-111 Change Listing This page lists the changes from 204733-110 to Revision 204733-111. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-111 FRONT Title Change Front Matter Content Control Number to 204734-111 Installation and Setup Changed Document Control Number to 204742-105 c4000 Installation ChecklistStandalone Under c4000 Install Accessories, changed Install Nameplate & Install Decal to Install Nameplate & Apply Decals. c4000 Install Accessories Deleted Install Cover and Cover Plate and added Install Nameplate & Apply Decals with two graphics 7A_9125A & 7A_9126a. 8 Revision ci4100 Installation Under ci4100 Integration, changed Install Nameplate & Install Decal to Install Nameplate & Apply Checklist-New Decals. c4000 and New i1000 ci4100 Integration Deleted Install Cover and Cover Plate and added Install Nameplate & Apply Decals with two graphics 7A_9125A & 7A_9126a. ci4100 Installation Under ci4100 Integration, changed Install Nameplate & Install Decal to Install Nameplate & Apply Checklist-New Decals. c4000 and New i1000 ci4100 Integration Deleted Install Cover and Cover Plate and added Install Nameplate & Apply Decals with two graphics 7A_9125A & 7A_9126a. ARCHITECT c4000 Service and Support Manual (Version 204733-111) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-112 Change Listing This page lists the changes from 204733-111 to Revision 204733-112. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-112 FRONT Title Change Front Matter Content Control Number to 204734-112 Troubleshooting Changed Document Control Number to 204737-107 AC/DC Controller Updated S1 & S2 Switch Setting table for S2-8 add c4000 and change setting to ON. MECH LED Updated LED number 9-10 to correct Function title error, changed to Reagent Bottle Sensor. 2 Revision ARCHITECT c4000 Service and Support Manual (Version 204733-112) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-113 Change Listing This page lists the changes from 204733-112 to Revision 204733-113. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-113 FRONT Title Change Front Matter Content Control Number to 204734-113 Troubleshooting Changed Document Control Number to 204737-108 Thermal Bath Controller (TBS) Board Adding New Digital TBC board and LEDs, Switches and Test pins information. 2 Revision ARCHITECT c4000 Service and Support Manual (Version 204733-113) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-114 Change Listing This page lists the changes from 204733-113 to Revision 204733-114. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-114 FRONT Title Change Front Matter Content Control Number to 204734-114 Planned Maintenance Changed Document Control Number from "204740-103" to "204740-104". 6 Revision Removed Planned Maintenance Procedures. Updated with ISA 128-006 ARCHITECT c4000 Service and Support Manual (Version 204733-114) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-115 Change Listing This page lists the changes from 204733-114 to Revision 204733-115. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-115 FRONT Title Change Front Matter Content Control Number to 204734-115 1 General Data Added attention activator under misc. symbol. 5 Procedures Added P-672, P-673, and P-674. Intallation & Setup 8 Changed Document Control Number from "204742-107" to "204742-108". Intallation & Setup 8 Updated installation procedures - Added new black articulated arm instruction to C4000 and i1000sr. Added Installation Checklist verification steps, attention activators. Revision ARCHITECT c4000 Service and Support Manual (Version 204733-115) • © 2009, 2012 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-116 Change Listing This page lists the changes from 204733-115 to Revision 204733-116. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-116 FRONT Title Change Front Matter Content Control Number to 204734-116 Troubleshooting Changed Document Control Number from "204737-108" to "204737-109". 2 Revision Updated DAQ BD Front Panel table. ARCHITECT c4000 Service and Support Manual (Version 204733-116) • © 2009, 2012 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-117 Change Listing This page lists the changes from 204733-116 to Revision 204733-117. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-117 FRONT Title Change Front Matter Content Control Number to 204734-117 Installation Changed Document Control Number from "204742-107" to "204742-108". 8 Revision Update/Correct unpack instructions and checklists. ARCHITECT c4000 Service and Support Manual (Version 204733-117) • © 2009, 2012 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-118 Change Listing This page lists the changes from 204733-117 to Revision 204733-118. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number to 204733-118 FRONT Title Change Front Matter Content Control Number to 204734-118 Installation Changed Document Control Number from "204742-108" to "204742-109". 8 Revision Update/Correct unpack instructions and checklists. ARCHITECT c4000 Service and Support Manual (Version 204733-118) • © 2009, 2012 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-119 Change Listing This page lists the changes from 204733-118 to Revision 204733-119. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-118 to 204733-119 FRONT Title Change Front Matter Content Control Number from 204734-118 to 204734-119 Installation Changed Document Control Number from "204742-109" to "204742-110". 8 Revision Update Installation Checklist to include link to the new procedures and steps to perform procedures. ARCHITECT c4000 Service and Support Manual (Version 204733-119) • © 2009, 2013 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-120 Change Listing This page lists the changes from 204733-119 to Revision 204733-120. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-119 to Revision 204733-120. FRONT Title Change Front Matter Content Control Number 204733-119 to Revision 204733-120. 8 Installation Change Document Control Number 204742-110 to 204742-111. Installation Add M&D 3526 Check Water Bath Tempature Revision ARCHITECT c4000 Service and Support Manual (Version 204733-120) • © 2009, 2013 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-121 Change Listing This page lists the changes from 204733-120 to Revision 204733-121. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-120 to Revision 204733-121. FRONT Title Change Front Matter Content Control Number 204733-120 to Revision 204733-121. 8 Installation Revision Change Document Control Number 204742-111 to 204742-112. Pressure Monitoring/Clot Detection Subsystem Installation ARCHITECT c4000 Service and Support Manual (Version 204733-122) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-122 Change Listing This page lists the changes from 204733-121 to Revision 204733-122. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-121 to Revision 204733-122. FRONT Title Change Front Matter Content Control Number 204734-121 to Revision 204734-122. 2 Troubleshooting Revision Change Document Control Number 204737-109 to 204737-110. Add New Style MECH LED Board Diagram with LED Reference ARCHITECT c4000 Service and Support Manual (Version 204733-122) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-123 Change Listing This page lists the changes from 204733-122 to Revision 204733-123. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-122 to Revision 204733-123. FRONT Title Change Front Matter Content Control Number 204734-122 to Revision 204734-123. 5 Revision General Data Change General Data Control Number 204736-108 to revision 204736-109 Procedures Added Remove Universal Shutter From Mount Plate Action to C1.02 Optics Shutter Added Install Universal Shutter on Mount Plate Action to C1.02 Optics Shutter Added Install Universal Shutter on Mounting Post Action to C1.02 Optics Shutter ARCHITECT c4000 Service and Support Manual (Version 204733-123) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-124 Change Listing This page lists the changes from 204733-123 to Revision 204733-124. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-123 to Revision 204733-124. FRONT Title Change Front Matter Content Control Number 204734-123 to Revision 204734-124. 8 Installation & Setup 8 Revision Change Control Number 204742-112 Revision 204742-113. Added Leveling instructions to the c4000 Standalone Installation Instructions ARCHITECT c4000 Service and Support Manual (Version 204733-124) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-125 Change Listing This page lists the changes from 204733-124 to Revision 204733-125. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from 204733-124 to Revision 204733-125. FRONT Title Change Front Matter Content Control Number 204734-124 to Revision 204734-125. 8 Installation & Setup 8 Revision Change Control Number 204742-113 Revision 204742-114. Installation checklist, pressure monitor PCB information. ARCHITECT c4000 Service and Support Manual (Version 204733-125) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. c4000 Revision 204733-126 Change Listing This page lists the changes from 204733-125 to Revision 204733-126. The revisions to the manual are indicated below. Be sure the manual contains the latest revision number of all pages. Section Numbers Sections Revised / Added FRONT Title Change Manual Revision Number from “204733-125” to” 204733-126”. FRONT Title Change Front Matter Content Control Number “204734-125” to “204734-126”. Chapter 8 Installation Change Control Number “204742-114” to “204742-115”. Revision Add note to c4000 Standalone ci4000; New c4000, New i1000SR, New c4000 Existing i1000SR, Existing c4000, New i1000SR and Existing c4000, Existing i1000SR Installation Checklists to verify module serial numbers (S/N) configured in the SCC with S/N on instruments. ARCHITECT c4000 Service and Support Manual (Version 204733-126) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Chapter 1 General Data (Document Control Number 204736-115) ARCHITECT c4000 Service and Support Manual (Version 204733-123) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. How to Use This Manual Links Biological Hazards Chemical Hazards Electrical Hazards Electrostatic Discharge (ESD) Hazards Laser Light Overview Safety Icons and Hazard Symbols Waste Handling and Disposal Overview The purpose of this manual is to provide information useful for servicing the ARCHITECT system and is composed of the following sections: Section Topic General Data This section contains a product overview, information on manual usage, accident prevention symbols and system specifications. Troubleshooting This section contains reference and normal operation information for the system. The information includes block and functional diagrams, mechanism and motor listings and LED operating conditions. Parts Lists This section provides lists of Supplemental Tools and Supplies, Customer and Field Service replaceable parts which are indexed by item number and description. Removal & Replacement This section contains Removal & Replacement procedures which are indexed by number to the Parts Lists. Procedures This section contains configuration, adjustment, calibration, checks and test procedures used to setup or verify instrument operation. Procedures are also used to assist in troubleshooting. Planned Maintenance This section contains preventative maintenance (PM) recommendation checklists. These recommendations may be subject to change due to local business or regulatory needs. Pre Site Specification & Checklist This section includes the site specifications and a checklist that outlines the activities of the Field Service Engineer/ Representative (FSE/FSR). An open action item summary is included in the checklist. Installation This section outlines the activities required to install or relocate a system. Hazards Introduction The c4000 is designed for optimal operator safety; however, this does not reduce the importance of safety awareness where hazards exist. This section describes the types and locations of potential hazards that could cause physical harm or damage to the laboratory environment, or where failure to follow instructions may result in instrument failure or generation of erroneous patient results. Warnings are inserted throughout this manual to alert Field Service Representatives (FSRs) to potential hazards. Hazard Signal Words Standard warning conventions, including hazard signal words and symbols, are described below. Signal Word DANGER! Definition Denotes an immediate hazard which, if not avoided, could result in serious injury or death. This signal word represents the highest level of any hazardous situation. Denotes a hazard which could result in moderate to serious personal injury. WARNING Denotes potential hazards that could result in minor injury. Also used for conditions or activities that could interfere with proper functioning or performance of the instrument. Caution Denotes operator or service information. Note Safety Icons and Hazard Symbols Safety icons in this manual and on the c4000 identify potentially dangerous conditions. Field Service Representatives must recognize the icons and understand the type and degree of potential hazard. The following icons may be used with text or in lieu of text. If text accompanies the icon, it describes the nature of the hazard and is labeled with DANGER, WARNING or CAUTION. In some situations, instrument labels refer Field Service Representatives to the manual for specific information. Symbol Hazard Caution Moving Parts Caution Chemical Hazard WARNING Splash / Spray Hazard WARNING Potential Biohazard WARNING Probe Stick Hazard WARNING Electrical Shock Hazard DANGER! DANGER! High Voltage High Voltage Caution Possible Electrostatic Discharge Shock Note Card Cage Ground Caution Class 2 Laser Radiation Description Identifies an activity or area where moving parts are present. Identifies an activity or area where hazardous chemicals are present. Refer to the Material Safety Data Sheet (MSDS) or package insert for specific safety information. Identifies an area where fluids may be under pressure. Identifies an activity or area where potentially infectious materials may be present. Follow procedures outlined in Biological Hazards. Identifies an activity or area where probes may be present. Identifies the possibility of electrical shock if procedural or engineering controls are not observed. Identifies high voltage areas over 600 volts. Identifies high voltage areas. Alerts the user to the possibility of electrical shock in noted activity or at posted location in the power supply. Follow precautionary measures outlined in Electrical Safety. Identifies an area where electrostatic discharge may be present. A ground strap must be worn while servicing the system. Identifies the location on the card cage where the ground strap is clipped. Warns against direct viewing into the bar code laser beam or reflections from the beam. Caution Hot Surface Caution Lifting Hazard Identifies an area where a hot surface is present. Identifies an activity where lifting or moving a heavy object requires assistance and/or appropriate lifting devices. The labeling of c4000 reagents/calibrators/controls or liquid consumables may include one or more of the following hazard symbols. The symbols and/or other country-specific warnings are used to convey properties of the chemical or chemical mixture and to notify the user that precautions should be taken when handling the material. Always consult the specific package insert or Material Safety Data Sheet for further information. Hazard Symbol Definition / Description (with Standard Abbreviation) Indicates the material is Oxidizing (O). Indicates the material is Highly Flammable (F) or Extremely Flammable (F +). Indicates the material is Toxic (T) or Very Toxic (T +). Indicates the material is Harmful (Xn) or an Irritant (Xi). Indicates the material is Corrosive (C). Indicates the material is Dangerous for the environment (N). Electrical Hazards The c4000 does not pose uncommon electrical hazards if installed properly and operated without alteration and connected to a power source that meets required specifications. Refer to Electrical, Power, and Heat Output Specifications in System Specification for details. Basic electrical hazard awareness is essential to the safe operation of any system. Only qualified personnel should perform electrical servicing. Elements of electrical safety include, but are not limited to the following: • • • • • • • • • • • • • Inspect electrical cabling into and on the c4000 periodically for signs of wear and damage. Turn the instrument OFF before disconnecting the power cord and before servicing any electrical or internal components. Determine the cause of a blown fuse or thrown circuit breaker and correct the problem before attempting to resume operation of the equipment. Only use replacement fuses of the specified type and electrical rating. Verify the power to the instrument is turned OFF. Be aware that a high voltage charge may remain on the power supply with the power OFF. Use an electrically insulated tool to disconnect the power supply and short both male pins to the instrument chassis. Keep liquids away from all connectors of electrical or communication components. Unplug the instrument before clean up of major liquid spills. Make sure hands are dry prior to touching any switches or outlets. Keep the floor dry and clean under and around the c4000. Use only approved power cords and electrical accessories, such as those supplied with the instrument, to protect against electric shock. Connect power cords only to properly grounded outlets. It is recommended that a ground fault circuit interrupter be used when working in a wet environment. Mechanical Hazards The c4000 is an automated system that operates under computer control. As with most automated equipment, there is potential for injury and bodily harm from moving mechanical components whenever the instrument is in operation. The c4000 minimizes mechanical hazards by providing guards to protect against accidental contact with moving components. The c4000 requires accurate positioning of all samples, reagents, calibrators, controls and consumables before executing any program. If safeguards must be removed or disconnected during setup and testing procedures, use caution. Stay a safe distance away from the moving components: fast-moving parts may cause injury. Always replace the safeguards upon completion of setup and testing procedures. It is never acceptable for instrument operators to reach into the instrument when it is in an operating mode. Should intervention be necessary during a run, the run should be interrupted according to instructions defined in the operations manual. Field Service Representatives of the c4000 may be potentially exposed to mechanical hazards when working on the following functional groups or subsystems: • C2 Reaction Carousel • D1 Reagent Supply Center • D2 Reagent Supply Center 2 • E4 Wash Solution Pump • E5 ICT Reference Pump • E6 ICT Aspiration Pump • E7 Inner Probe Wash Pump Bellows • E8 H.C. Waste Pump Peristaltic • E9 Cuvette Wash Pump, Bellows • G1 Sample Syringe Drive • G2 Reagent Syringe Drive • R1 Mixer Assembly • R2 Sample Pipettor • R3 Sample Pipettor Head • R4 Reagent Pipettor Elevator • R5 Reagent Pipettor Head • R6 Cuvette Washer Assembly • R7 ICT Assembly • S1 Sample Carousel • Robotic Sample Handler (RSH) Injury may result in the event of an instrument malfunction or an unexpected sequence of mechanical component movements. Basic elements of mechanical equipment safety include but are not limited to: • • • • • • • • Never bypass or override a safety device unless during setup or testing activities Never operate the instrument without protective covers and barriers in place unless during setup or testing activities Never perform manual tasks on the work surface of the c4000 Never allow any part of body to enter a range of mechanical movement during system operation Do not wear articles of clothing or accessories that could catch on the c4000 Keep pockets free of items that could fall into the c4000 Be especially cautious when performing adjustment, maintenance, cleaning, or repair procedures Use caution when loading sample carriers and reagents. Laser Light The c4000 is a Class 1 laser product containing embedded Class 2 lasers. When this product is used according to operating instructions, the laser does not present a hazard to the eyes. Failure to follow proper procedure may result in an eye injury. • • • • Only Abbott trained Field Service Representatives should remove the inner protective covers that enclose the embedded laser. Do not remove, damage or obliterate any of the laser warning labels. Never stare directly into the beam, look into the aperture, place any optics into the beam or bypass interlocks. Replace any label that becomes illegible. Laser Caution Labels On the RSH (Robotic Sample Handler), the following label is affixed above the RSH bar code reader under the transparent, priority bay cover. On the c4000 and i1000SR, the following label is affixed to the inside of the right door on the processing module. Note The i1000SR is certified to IEC 60825-1 (2001). Biological Hazards The following activities may involve the presence of potentially infectious materials: • • • • • • • Handling samples, reagents, calibrators controls Cleaning spills Handling and disposing of waste Moving the system Performing maintenance procedures Performing decontamination procedures Performing component replacement procedures Precautions All instruments used to process clinical specimens, reagents, controls, calibrators, etc., that contain human-sourced material must be treated as potentially infectious. Follow ADD's Biosafety Procedures, which include, but not limited to, the following precautions: • Wear gloves, lab coats and protective eyewear when handling human-sourced material or contaminated instrument components. • Do not pipet by mouth. • Do not eat, drink, smoke, apply cosmetics, or handle contact lenses when handling human-sourced material or contaminated instrument components. • Clean spills of potentially infectious materials and contaminated instrument components with a detergent followed by an appropriate disinfectant, such as 0.1% sodium hypochlorite or other suitable disinfectant. • Decontaminate and dispose of all potentially contaminated materials into the appropriate biohazard waste system in the facility. • Dispose of probes and other sharp objects in appropriately labeled, puncture-resistant and leak-proof containers. Note Refer to EHS Policies/Guidance BioSafety If exposed to infectious or potentially infectious materials, immediately cleanse the affected area, then seek medical attention as soon as possible: • • • • Eyes rinse with water for 15 minutes Mouth rinse with water Skin wash the affected area with soap and water Puncture wound allow to bleed freely. Wash the affected area with soap and water Note Report all accidents to your Manager and Workers' Compensation Administrator. For additional information see Site Documentation: - ADDDA_SAF0201 GCO EHS Incident Reporting - BD30-004 Incident Reporting and Investigation Spill Clean up Clean spills in accordance with established biosafety practices and follow instructions provided in the MSDS. In general, safe work practices for cleaning spills include: 1. Wear appropriate personal protective equipment such as gloves, lab coats and protective eyewear. 2. Absorb the spill with absorbent material. 3. Wipe the spill area with detergent solution. 4. Wipe the area with an appropriate disinfectant, such as, a 0.1% sodium hypochlorite (10% solution of chlorine bleach containing at least 5% sodium hypochlorite). 5. Dispose of spilled and contaminated material in accordance with the facility’s waste disposal procedures. Instrument or Part Decontamination Any part subject to user handling (for instance: keyboards, printers, monitors, disk drives, front panels, etc.) requires decontamination prior to servicing or shipment. When handling or decontaminating spare parts for shipping or repair, use safety precautions as listed below. Caution Always wear appropriate personal protective equipment (protective eyewear, gloves, lab coat) while performing decontamination activities. 1. Remove all specimens, reagents, controls, calibrators, etc. from the instrument. 2. Cycle sodium hypochlorite solution (chlorine bleach) through the fluid pathway (including probes, if applicable) that contacted the human-sourced material or products containing potentially infectious material. Allow the solution to stand for a minimum of 10 minutes. If a chlorine bleach solution cannot be used, cycle buffer or water through the fluid pathway to thoroughly flush the fluid pathway. 3. Remove all waste materials from the instrument, if present. 4. Rinse waste containers with 0.1% sodium hypochlorite or other suitable disinfectant prior to shipping or disposal. 5. Wipe down the surface of each instrument and component with a detergent solution followed by an appropriate disinfectant, such as 0.1% sodium hypochlorite. Note Under normal circumstances, printed circuit boards do not require decontamination. Field Replaceable Units (FRUs) enclosed inside the skins of computer and peripheral equipment are not considered contaminated. Decontamination may affect the performance of a printed circuit board or internal computer component. Waste Handling and Disposal It is the responsibility of each facility to label all waste containers and to characterize its waste stream to ensure the waste is disposed in accordance with appropriate waste disposal regulations. Dispose of sharps in an appropriately labeled, puncture resistant and leakproof container. Products That Contain Thimerosal or Mercury Some reagents, calibrators and controls contain thimerosal or mercury and may be considered hazardous by various environmental regulatory agencies. Do not directly sewer waste fluids that contain thimerosal or mercury. Products That Contain Guanidine (Thiocyanate) Do not mix oxiding agents, such as sodium hypochlorite (chlorine bleach), with liquid waste. Toxic gases may be generated. Chemical Hazards Field Service Representatives may be exposed to hazardous chemicals when handling reagents, calibrators, controls, or liquid consumables (including bleach). Exposure to hazardous chemicals is minimized by following instructions provided in the assay Package Inserts and Material Safety Data Sheets (MSDS). Exposure levels are further reduced by the design features of the instrument when it is used properly. Precautions In general, observe the following precautions when handling chemicals: • Consult Material Safety Data Sheets for safe use instructions and precautions. • Avoid contact with skin and eyes. If contact with material is anticipated, wear impervious gloves and protective eyewear and clothing. • Always maintain good housekeeping. Do not eat, drink or store food and beverages in areas where chemicals are used. • If irritation or signs of toxicity occur after exposure, seek medical attention. Hazard symbols that appear on c4000 product labeling are accompanied by Risk (R) and Safety (S) numbers representing specific risk and safety phrases defined by applicable European Community Directives. The risk and safety phrases describe precautions used when working with a particular chemical or chemical mixture. Refer to the corresponding phrases indicated in the Package Insert or similar document for all (R) and (S) numbers that appear on product labeling. Other country-specific warning and precautions may be included on the labeling. Sodium Azide Some products contain sodium azide. Observe the following precautions when using products that contain sodium azide: • Do not mix any chemical or product with a pH below 6 with solutions or residues that contain sodium azide (such as wash buffer). Sodium azide releases hydrazoic acid, a very toxic and pungent-smelling gas when the pH is lower than 6. CautionThe pre-trigger solution has a pH lower than 6. Allowing the pre-trigger solution to contact wash buffer or dried wash buffer residue may result in the release of hydrazoic acid, a very toxic and pungent-smelling gas. Care must be taken to clean up any spills or residue that may have occurred due to leaking components (ex. pumps in the pump bay) identified during servicing of the instrument. • Flush drains thoroughly with water several times a day to prevent potentially explosive metal azides from forming on lead, copper or brass components or on solder in laboratory plumbing if product and/or instrument waste is released to a drain. Detailed information about azides in laboratory drains is available in Current Intelligence Bulletin No. 13 Explosive Azide Hazard (August 16, 1976), a publication issued by the U.S. National Institute of Occupational Safety and Health (NIOSH), at cdc.gov/niosh. A copy of the bulletin is also available on the Abbott Diagnostics Global internet site at abbottdiagnostics.com (select the International region), under Support - Technical Library - Other Reference Documents, or by contacting your Abbott representative. Physical Hazards Safe practices should be observed while working with the following physical hazards to avoid potential injuries. Sharps and Probes In general, the use of sharps and glassware should be minimized. The probes are sharp and may be contaminated with infectious materials. Avoid contact with these components and handle them cautiously in order to prevent injury. Use mechanical means to remove contaminated broken glassware. When handling the external waste pump (c4000 optional accessory), use proper personal protective equipment. Dispose of sharps in an appropriately marked, puncture resistant and leakproof container before treatment and disposal. Heavy Objects The system is heavy and sections of the shell are unsupported. Ensure you have adequate help before attempting to move the system. Push only on solid sections of the housing; do not exert pressure on unsupported sections of the shell. The c System high-concentration waste container is heavy when full. The i System wash buffer reservoir is heavy and bulky when full. Use care when handling the container to reduce the risk of injury. When moving or lifting heavy objects, obtain assistance and always use proper lifting techniques. Hot Objects The buffer heaters, lamp and lamp housing may be hot. Allow the surface to cool before touching and handling. Use temperature resistant gloves, if necessary. Trip Hazard The c4000 is equipped with a power cord and various computer connectors. To avoid a tripping hazard, ensure cords in high traffic areas are properly stowed. Electrostatic Discharge (ESD) Many of the electronic components on the c4000 circuit boards are susceptible to electrostatic discharge (ESD). Static discharge of as little as 100–200 volts can damage or destroy a component. Always wear a wrist ground strap and discharge static electricity from your body prior to touching and working on the instrument. Note All card cage boards are shipped with a ground strap and graphical instructions for use. This symbol is placed on the card cage to identify the appropriate ground location. Static Hazard Static protective procedures are used during the manufacture of PC boards. Replacement PC board assemblies are also protected by use of static protective packaging as well as boxed to prevent physical damage. Assemblies that have failed and are returned for repair are also handled at the repair shop under static protection procedures. Handling Guidelines PC Subassemblies These guidelines assure protection against failures created by static: • Retain spare PC board subassemblies in the static protective bags. • Use an approved static protective field service kit, or the ground strap shipped with the board, whenever a board is removed from an instrument or protective bag. • Replace the defective PC board in the same protective bag to return for repair. Continued use of the protective shipping boxes, both during shipping and storage, eliminates most failures caused by physical damage. Static Protective Service Kit The static protective service kit is designed to keep the FSE/FSR, replacement part, work surface and instrument at the same ground level. Generally, an instruction set accompanies the kit. In the absence of specific instructions, follow ESD Procedure. ESD Procedure Note Use where ESD symbol is present and static protective equipment is not shipped with replacement part. 1. Place the work mat on a solid surface close to the instrument. Allow the ground strap to reach the instrument. 2. Attach the ground clip to the instrument chassis. 3. Attach the other end of the ground clip cable and the connector from the wrist strap to the work mat. (Exception: Some wrist strap cables provide a clip to connect to the same ground source as the mat cable). 4. Attach wrist strap to your wrist; make sure the metallic button on the inside of the wrist strap is in direct contact with skin. 5. Place PC boards removed from the instrument on the work mat. 6. Replacement PC boards should be placed on the work mat before removing from the protective bag and remain on the mat until installation. 7. Defective PC boards should be placed in the static protective bag before removal from the work mat area. Miscellaneous Symbols Attention Activator Kit Parts Static protective work surface Wrist strap and attaching cable Grounding clip or cable ARCHITECT c4000 Service and Support Manual (Version 204733-115) • © 2009, 2012 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. c4000 Module Overview Links c4000 Processing Module Fluidics Integrated Chip Technology (ICT) Liquid Level Sense (LLS) PCB Photometric Technology Power Supply Temperature Control System Valves, Motors and Sensors c4000 Processing Module Overview The ARCHITECT c4000 is a high performance clinical chemistry diagnostic system capable of quantifying selected analytes in biological fluids. The system consists of the c4000 Processing Module, the Robotic Sample Handler (RSH), and a separate System Control Center (SCC) for initiating and monitoring operations. The system can be configured with up to 220 different assays including Ion Selective Electrode (ISE) analysis of sodium, potassium, and chloride with up to 90 different cartridges onboard. It can process samples at 400 tests/hour without ICTs and a maximum throughput of 600 tests/hour when running both photometric and ICT tests (9 second system cycle). Individual assay time is a maximum of 10.5 minutes for photometric tests and 4 minutes for ICT tests. Operational modes include random, continuous access and priority processing. Endpoint and rate reactions are calculated using absorbance measurements that are converted into concentration units. Absorbance measurements are made using a diode array detection system. The Integrated Chip Technology (ICT) for electrolyte determinations consists of an integrated module using ion selective electrode technology. Quality control (using Levy-Jennings and Westgard rules) and on-board data storage is located on the hard drive of the SCC. The c4000 reaction carousel contains 99 permanent, rectangular cuvettes in a temperature controlled water bath maintained at 37°C for a stable reaction process. The cuvettes are held in 9 cuvette holders (segments) designed for the c4000 system. Each of the 9 segments contain 11 cuvettes. The reaction carousel rotates clockwise 1 ¼ turn every nine seconds during processing and positions the cuvettes for sample and reagent dispensing, mixing, photometry and cuvette washing. Each individual cuvette is read as it passes through the optical light path. A water distribution system is used to wash the cuvettes, sample probe, reagent probes and mixer blades. The cuvette washer is used to distribute diluted alkaline and acid solutions during the cuvette wash process. The mixer assembly utilizes 2 mixers to mix the sample and reagent in preparation for final analysis. The c4000 has 1 reagent supply center that contains both the R1 and R2 reagents. The inner and outer carousels are split into 15 segments and have a total of 90 positions for 55/20 mL size wedge bottles, or 66 positions for 90 mL size wedge bottles. The RSH is used to move the carriers to the proper position for sampling. The RSH has total 20 slots for loading/unloading sample and carriers, and also has 3 pockets on the carrier positioner for continuous sampling. Pressure monitoring circuitry is incorporated for clot and bubble detection on the sample pipettor. An alert is issued when a clot or bubble is detected providing the operator the opportunity to rerun or check the quality of the sample. Pressure monitoring is also provided for the R1 and R2 pipettors. Assay Processing Sequence Position(s) Description 1 The sample pipettor dispenses sample into the cuvette at the starting position. 2 The reaction carousel rotates approximately 1 and ¼ turns (124 cuvette positions). The cuvette containing sample is now at the first reagent dispense position. Reagent pipettor 1 dispenses reagent 1 in the cuvette. 3 The reaction carousel rotates on cycle (1 and ¼ turns) to the first mixing position where the mixer unit (mixer 1) mixes the sample and reagent 1. Note Each time the reaction carousel rotates, the cuvette passes the photometric position where the lamp is located and the photometer measures the absorbance. 4 The reaction carousel rotates one cycle to this position at which no activity takes place. 5 The reaction carousel has completed 4 cycles. Each cycle rotates the reaction carousel 124 cuvette positions (approximately 1 and ¼ turns). Since there are 99 cuvettes in the reaction carousel, each cycle indexes the carousel 25 cuvette positions (124-99=25). After 4 cycles the cuvette is now one position beyond the original starting position (4x25=100). 6-68 The reaction carousel continues to rotate and the reaction mixture incubates. The photometer takes absorbance readings every time the cuvette passes the photometric position. Although the cuvette passes the photometric position 68 times, the c4000 processing module only uses up to 33 absorbance readings for calculation of results. 69-97 The cuvette washer removes the reaction mixture to waste and cleans the cuvette with Alkaline Wash, Acid Wash, and DI water. Then the cuvette washer dispenses DI water into the cuvette for a water blank measurement to ensure cuvette integrity. Finally, the cuvette washer aspirates the water and dries the cuvette. 100 The clean cuvette waits for new sample dispense. Some assay protocols may also use the following locations: Position(s) Description 5 If onboard dilution is required, the sample pipettor aspirates the diluted sample and dispenses the sample into the new cuvette that is currently at position 1. 22 For an ICT sample, the ICT probe aspirates the diluted sample into the ICT unit. 37 If the reaction requires a second reagent, reagent pipettor 2 dispenses reagent 2 into the cuvette. 38 The mixer unit (mixer 2) mixes the second reagent with the sample and reagent mixture. Top View Front View Rear View Side View Photometric Technology Photometry is defined as the measurement of light, which travels in waves. The color of light is a function of its wavelength, which is the distance between the peaks of a light wave. As the wavelength of light is varied, a visible change in color is detected. Material that appears to have color absorbs light at specific wavelengths and reflects others producing different shades of color. For example, a substance that absorbs blue light at 475 nm reflects all other wavelengths and is seen as yellow, which is the sum of all other reflected light. In a laboratory setting an assay specific reagent solution is mixed with a patient sample causing a chemical reaction and color change to take place over time. Photometric technology uses a light source and optics unit to measure the kind and the amount of light absorbed or transmitted by the solution. This technology follows Beer's Law, which states the absorbance of a solution is directly related to its concentration. The relationship between transmittance and absorbance as a function of concentration is explained as follows: Transmittance is the amount of light that passes through a solution and is measured photometrically. The amount of light reaching the optics is expressed as the percent transmittance. As the concentration of the color in the sample is increased, the percent transmittance is decreased although the relationship is not linear. A more linear relationship is established by plotting the logarithm of the %T. Absorbance measures the amount of light that is absorbed by a solution. Absorbance describes the percentage of light transmission and is directly proportional to concentration. Standards or calibrators with known concentration values are measured in a photometric system. The measured absorbance values are plotted against the known concentrations of the calibrators and a calibration curve is drawn. The absorbance of the patient sample is measured and compared to the calibration curve and the concentration value of the patient result is calculated. End-Point Assay End-point assays are reactions that reach equilibrium and at that time, there is little or no further change to the absorbance readings. During this equilibrium time, the system measures the absorbance readings used for calibration and calculating results. For end-point assays, the system calculates the concentration using the absorbance data obtained in the Main Read Time specified on the Configure assay parameters window - General Reaction Definition view. End-Point Assay Reaction Curve Example Rate Assay Rate assays are reactions that measure the constant change in absorbance over time. The system performs readings several times during this reaction, calculates absorbance change (activity), and uses the readings to calculate results.For rate assays, the system calculates activity using the change of absorbance per minute in the Main Read Time specified on the Configure assay parameters window - General Reaction Definition View. Optics System Serum or urine sample is dispensed into a glass cuvette and mixed with reagent. In the constant reaction process, variable rate of absorbance or final absorbance at the end of the reaction process is used for calculating the measurement result. A spectrophotometer is used to monitor the absorbance at 16 defined wavelength(s). The 99 cuvettes in the reaction carousel rotate in a water bath that maintains constant temperature (37°C) for a stable reaction process. The optical system of the c4000 module is a reaction cuvette direct photometry system that serves to direct and align the light from the source lamp, through the water bath and cuvette, and into the optics unit. The system focuses only light that originates at the source lamp. The analyzer uses a concave diffraction grating and is capable of taking monochromatic and bi-chromatic measurements at 16 different wavelengths. Monochromatic measurements compare the absorbance of a chromogen in a sample solution against a water or reagent blank. Bi-chromatic analysis uses the absorbance difference of a chromogen between two pre-selected wavelengths. C4000 Optics Assembly Light from the tungsten halogen source lamp passes through the aperture, heat absorbing filter, convex focusing lens, water bath lenses and cuvette. The heat absorbing filter protects the first convex lens. The light passes into the optics assembly through another convex focusing lens and is reflected off of a mirror 90 degrees through a slit to a diffraction grating. The slit controls the size of the light beam onto the grating assembly. The grating breaks up the focused light beam into the 16 component wavelengths (340nm to 804nm) and reflects the light spectrum to the diode array. The photo diode detects and measures changes in light intensity. The current signal from the photo diode array is changed to a voltage and is amplified by the pre-amp board. In the DAQ board circuitry, the signal is inverted and selected from 16 channels by a multiplexer. The Trigger Sensor signals the DAQ board the cuvette is in the light path and is in position for the readings to take place. A gain control circuit (on the DAQ board) changes the signal to an optimum level, converts the analog signal to a digital value and passes the signal to the CPU board. Final result calculations are performed by the system control center. DAQ Beginning of a Sample Processing Run • Photometric gains are set to compensate for lamp degradation: - Light passes through cuvette #34and into the optics assembly where the Diffraction Grating separates the light into 16 different wavelengths and focuses them onto 16 different photo-diodes. The current signals from the photo-diodes are changed to a voltage equivalent and amplified by the pre-amp board in the optics assembly. The signal range from the output of the pre-amp is 0V (dark) to -5V max - The DAQ Board inverts the signal (0V to +5V) and multiplexes it to the DAQ gain control circuitry - The DAQ gain control circuitry increases or decreases the signal level input to the DAQ ADC circuitry. For example, if the output of the pre-amp is –1.5V, a gain of x2 is selected and the input level to the ADC becomes 3V - The amplified signals are converted by the DAQ to a digital value of 35000 (2.5VDC) to 65000 (5.0VDC) with a maximum gain of 4.0. • The optics shutter is closed, and a dark current read is taken. This value is used to offset noise in the detection circuitry and is used in the absorbance calculation • Cuvette #34, containing water, is used to perform a trigger check. During the trigger check, the system rotates the reaction carousel and locates the cuvette segment tab for the cuvette. After the trigger sensor detects the trailing edge of the tab, the optics are used to calculate the center of the cuvette by locating both edges of the cuvette. The difference from the target (expected) value is the offset. The offset value is used to determine when the system begins the optical readings on the individual cuvettes as they pass by the optics • The system performs a water blank/cuvette integrity check optical read on each cuvette. The digital readings from the DAQ are sent to the SCC via the CPU and are used to check the individual cuvette integrity. If the check passes, the readings are used as the water blank value in the absorbance calculation. If it fails the system skips the cuvette during the run and generates an error message. During a Sample Processing Run • Sample is dispensed into a cuvette, reagent is added and mixed • Thirty-three reads are made at the sixteen wavelengths. Note The c4000 module reads a total of 68 points of photometric data. However, the SCC utilizes 33 read points to keep equal conditions throughout “C” series instruments. • The digital readings from the DAQ are sent to the SCC via the CPU and depending on the assay reaction type and configuration, the software extracts the proper wavelength readings for the selected read time • The unknown sample value is used in the absorbance calculation. The measured absorbance values are plotted against the known concentrations of the calibrators and a calibration curve is drawn. The absorbance of the patient sample is measured and compared to the stored calibration curve and the concentration value of the patient result is calculated. Abs = -10/5log ((sample voltage – dark voltage) / (water blank voltage – dark voltage)) Lamp Monitoring This function monitors the lamp energy, to detect lamp output fluctuations. This minimizes generation of aberrant results caused by normal lamp degradation. • Lamp monitoring is performed during the cuvette water blank measurement • Reads are a bi-chromatic calculation of the absorbance at 340nm and 380nm wavelengths • Readings are analyzed against recent lamp data • A range (Max-Min.) of the bichromatic reads are calculated and compared to the predefined threshold in the system software parameters • Bichromatic optics errors are generated when the reads are outside of specifications due to lamp instability and/or an issue with the water bath Ending of a Sample Processing Run • The cuvette washer removes the reaction mixture to waste, and cleans the cuvette with Alkaline Wash, Acid Wash, and DI water • The cuvette washer dispenses DI water into the cuvette for the water blank measurement • The cuvette washer aspirates the water and dries the cuvette in preparation for the next sample • If additional sample analysis is not requested, the sample probe dispenses water into the cuvette for hydration Power and Card Cage Components Component Function and Description Lamp Power Supply • Located in the Main Power Supply • Provides an adjustable +11.45VDC lamp voltage DAQ Board • Located in Master Card Cage • Provides gain control circuitry for optical signals • Converts the analog optical signals from the pre-amp into a digital equivalent (A/D conversion) CPU • Located in Master Card Cage • Receives optical signal from DAQ board and sends them to SCC for result processing Optic Assembly and Lamp Housing Components Component Function and Description Optics Assembly Note There are no field adjustable, cleanable, or replaceable parts in the Optics Assembly. Shutter • Shutter is normally open • During RUNNING Status -closes for dark current and stray light reading Convex Lens • Focuses light onto mirror Diffraction Grating • Breaks the beam into 16 wavelengths and reflects the light onto a Photo Diode Array Photo Diode Array • Detects and measures changes in the light intensity and sends signal to the Preamp PCB Preamp PCB • Converts optical signal to a voltage, amplifies it and sends it to DAQ board Optical Fan • 100VAC cooling fan secured to outside of Optics Assembly Halogen Source Lamp • • • • • +11.45VDC from DC Lamp Power Supply located in Main Power Supply Lamp is always ON Filament should be vertically positioned Replaced per customer quarterly maintenance Located in center of Reaction Carousel Assembly Water Bath Components Component Water Bath Function and Description Heat Absorbing Filter Protects the Convex Lens from heat distortion Convex Lens Focuses light path onto Cuvette Water Bath Lens (2) and gaskets Allows light to pass through the water bath Cuvette Cuvettes in the reaction carousel rotate in a water bath that maintains constant temperature (37°C) for a stable reaction process. Cuvette Wipers Removes bubbles from the outside of the cuvette Integrated Chip Technology (ICT) The ICT module is an integrated subsystem for determining sodium, potassium and chloride in fluid samples. The ICT module contains a solid state ion selective electrode for each analyte. Each electrode is based on the unique properties of ion specific membranes that develop an electrical potential or voltage when exposed to a solution containing the analyte it is designed to measure. The voltage developed varies in response to concentration changes – the higher the sample concentration, the larger the voltage output. The voltage on the ion selective electrode is compared to a reference electrode that is an integral part of the ICT module. ICT Module and Flow Electrode Description Sodium (Na+) A Crown Ether ionophore incorporated into an ion-selective plastic membrane. (It is NOT a glass electrode, and therefore is less affected by pH changes) Potassium (K+) Valinomycin incorporated into an ion-selective plastic membrane. Chloride (Cl-) A solid silver chloride (AgCl) disk. Reference A silver/silver chloride electrode in a potassium chloride (KCl) gel inner solution, separated from the sample by a porous ceramic tube. The c4000 module uses an on-board bulk solution known as ICT reference solution. Reference solution is used in the ICT fluidic system in determination of electrolyte values. The solution serves to stabilize the temperature of the ICT electrodes, provides known concentration or standard for calculating the concentration of the sample and to rinse out the sample at the measurement process. Proper fluidic operation is critical for accurate and precise results. A capacitive sensor located in the reference solution cup monitors fluidic operation. If the cup does not fill or if the reference solution is not aspirated an error is generated. ICT Reference Solution Cup Level Sense Description The ICT Reference Solution Cup contains level sense electrodes used to sense the presence of ICT Reference Solution. When the cup is full, the solution covers both electrodes and provides a conductive pathway between the electrodes. After the ICT Module completes aspiration, the fluid level drops below the electrodes and there will be no conductive pathway between the electrodes. The analyzer confirms the ICT Reference Solution is dispensed and aspirated correctly by detecting the solution level in the ICT Reference Solution Cup before and after aspiration. If solution is not detected prior to aspiration, the systems generate error code 5622 Unable to process test, ICT reference solution cup not filling. If solution is detected after aspiration, the system generates error code 5623 Unable to process test, ICT reference solution not aspirated. When the ICT Reference Solution is not properly detected, the analyzer stops measuring the ICT assay, flags associated results and generates an error code that is added to the message history log. The analyzer stops measuring ICT Assays, but the cup is not drained. The level of solution left in the cup after an error is the same level of solution that generated the error. The level detect electronic signal pathway is as follows: 1. ICT Reference Solution Cup Sensors. 2. IREF Level Detect Board. 3. CNN. 4. CPU. IREF Level Detect Board LED Operation The IREF Level Detect Board is mounted on the side of the ICT Pre-amp board (located at the front, upper right of the c4000 module). There are two Light Emitting Diodes (LED) located on the board.LED 2 turns on only when the c4000 module CPU is reading the status of the level sense electrodes in the ICT Reference Cup. During this brief status check (when LED 2 is on) observe LED 1. If the cup is full during the status check, LED 1 will be on at the same time as LED 2. If the cup is empty, LED 1 will be off when LED 2 is on. Just before the ICT assembly moves down to aspirate, both LED2 and LED1 turn RED momentarily (cup is full), and turn off. After the ICT assembly aspirates and moves up, LED2 turns RED momentarily and turns off. LED1 never turns on because the IREF Cup is empty. Serum and Urine Sample Progression Schematic Diagram ICT Fluid Flow Diagram Link 7A_9001a.pdf Serum Sample Progression If concentrated ICT Diluent is used, the R1 pipettor aspirates 69uL of ICT Diluent and dispenses it into the cuvette along with 276uL of water. The ICT Reference Solution read in step four is used to condition the ICT and is not used in result calculations. 1. 15ul of sample is aspirated and dispensed into a cuvette by the Sample Pipettor. 2. The R1 Pipettor aspirates 345uL of ICT Diluent and dispenses it into the cuvette containing the sample. 3. The ICT Diluent and sample solution are mixed utilizing the R1 Mixer. 4. The ICT Assembly moves over the ICT Reference Solution Cup, aspirates ICT Reference Solution and reads for 9 seconds. (This step may occur multiple times if the system was in standby or if there was a pause between ICT runs.) 5. The ICT Assembly moves over the cuvette containing the sample and ICT Diluent mixture, aspirates and reads for 9 seconds. 6. The ICT Assembly moves over the ICT Reference Solution Cup, aspirates ICT Reference Solution and reads for 9 seconds. 7. The difference between the sample voltage reading made in step 5 and the ICT Reference Solution voltage reading made in step 6 are used to calculate the result. Urine Sample Progression The R1 Pipettor aspirates 345ul of ICT Diluent and dispenses it into the cuvette containing the sample. If concentrated ICT Diluent is used, the R1 pipettor aspirates 69uL of ICT Diluent and dispenses it into the cuvette along with 276uL of water. 1. 15ul of sample is aspirated and dispensed into a cuvette by the Sample Pipettor. 2. The ICT Diluent and sample solution are mixed utilizing the R1 Mixer. 3. The ICT Assembly moves over the ICT Reference Solution Cup, aspirates ICT Reference Solution and reads for 9 seconds. (This step may occur multiple times if the system was in standby or if there was a pause between ICT runs.) 4. The ICT Assembly moves over the cuvette containing the sample and ICT Diluent mixture, aspirates and reads for 9 seconds. 5. The ICT Assembly moves over the ICT Reference Solution Cup, aspirates IREF solution and reads for 9 seconds. 6. The ICT Assembly aspirates ICT Reference Solution a second time and reads for 9 seconds. 7. The difference between the sample voltage reading from step 5 and the ICT Reference Solution voltage reading made in step 7 are used to calculate the result. Note The ICT Reference Solution read in step four is used to condition the ICT and is not used in result calculations. The ICT Reference Solution read in step 6 is a flush used to eliminate potential carryover from a urine sample. Fluidics Fluidics Overview The Water Management Unit (WMU) provides incoming water distribution for the c4000 module. The WMU is located on the left rear side of the instrument. The unit consists of a number of valves, pumps and tubing that direct clean reagent grade water used for the water bath, probe wash, mixer washing and cuvette washing. On-board the WMU is a 3-liter internal water supply tank which fills automatically to maintain water availability and enables water to be distributed throughout the c4000 module when needed. Float switches in the WMU tell the c4000 module if there is adequate water available and if the low concentration waste drain system is working properly. A magnetic pump (MP), located in the WMU, pumps water to the sample probe, reagent probes and mixer wash cups for external probe washing. A water degasser is incorporated in the system to reduce bubbles in the DI water for the Probe Wash Pump, Cuvette Wash Pump, Wash Solution Pump, and Water Bath Fill Pump. • The bellows style Probe Wash Pump provides degassed water for dispensing and internal probe washing by directing the degassed water through the inside of the sample and reagent probes. • The bellows style Cuvette Wash Pump provides degassed water for washing the cuvettes by directing the degassed water through a manifold and to the various cuvette wash nozzles located on the cuvette washer head. • The syringe style Wash Solution Pump uses degassed water to create and deliver a 1.3% alkaline and 0.9% acid solution to the wash nozzles for cuvette washing. • The peristaltic style Water Bath Fill Pump provides degassed water for the reaction water bath. There are two syringe style pumps used in the ICT subsystem: • The syringe style ICT Aspiration Pump is used to aspirate sample and reference solution into the ICT module for analysis. • The syringe style ICT Reference Solution Pump is used to fill and drain the ICT Reference Cup with ICT Reference Solution. The cuvette washer generates high concentration waste (HCW) and low concentration waste (LCW) during the process of cleaning the cuvettes. The peristaltic style HCW Pump removes the high concentration waste from the cuvettes and sends it to the HCW waste drain. • Two diaphragm style vacuum pumps, VP1 and VP2, remove the low concentration waste from the cuvettes and send it to the LCW Tank where it is gravity fed to an external drain hose. The following schematic provides greater detail for the overall fluidics operation of the instrument: Water Management Unit (WMU) Schematic Diagram Fluidics Diagram Link 7A_9002b.pdf Water Management Unit (WMU) The Water Management Unit provides incoming water distribution and is located on the left rear side of the c4000 module. The WMU directs water to the water bath, probe wash cups, mixer wash cups and cuvette washer. 1 Water Flow Adjustment Valves 2 LCW Tank 3 LCW Float Switch 4 Incoming Water 1 Tank Float Switches 2 Water Supply Tank 3 Magnetic Pump On-board the WMU is a 3-liter internal water supply tank that fills automatically to maintain water availability and enables water to be distributed throughout the c4000 module when needed. Float switches in the WMU tell the c4000 module if there is adequate water available and if the low concentration waste drain system is working properly. A magnetic pump (MP), pumps water to the sample probe, reagent probes and mixer wash cups for external probe washing. WMU Input Specifications Water Quality C4000 Water Connection Flow Rate > 1 Meg Ohm-cm @ 25°C (77°F) Max Bacterial Content: 1000 colony forming units per mL (cfu/ml) Regulator and pressure gauge located prior to c4000 water input. No brass fittings in the water input line. Water pressure: 15-25psi at 15°C to 37°C Tubing: 3/8” (9.5mm) OD hard tubing Fitting: 3/8”-1/4”NPT (9.56.35mm) male thread (both ends) compression fitting The flow rate is determined by the water pressure and tubing size. If regulator pressure, tubing and connectors meet specification, the flow rate is acceptable. Consumption < 18L/hour WMU Tank Fill Float sensors inside the water tank tell the c4000 module when water is needed, when water is too low and when an overflow situation exists. Once the tank is filled, the remainder of the c4000 module can be supplied with water. 1 Float Sensors 1 Low 2 Full 3 Overflow The water inlet port fitting directs water through an on/off valve and flow restrictor that is preset from the factory. When water is needed, the incoming water valve is opened and the 3-liter open water tank fills with water through a flow restrictor. 1 Water Inlet 1 On/Off Valve 2 Flow Restrictor 3 Water Inlet WMU To Probe Wash Cups The magnetic pump (MP) pumps water from the water tank to the wash cup adjustment valves and solenoids. Each valve can be manually adjusted to distribute the correct amount of water to the sample probe wash cup, reagent probe wash cups and mixer wash cups. When water is not needed for probe and mixer washing, the valves remain open and water is re-circulated back into the water supply tank through a second manifold. The magnetic pump is constantly running to maintain circulation. 1 Mixer Wash Cups 2 Reagent Probes Wash Cups 3 Sample Probe Wash Cup 4 Water Flow Adjustment Valves Low Concentration Waste (LCW) Drain Manifold The LCW Drain Manifold is located on the left side of the c4000 module. It collects the waste from the sample probe wash cup, reagent probe wash cups, mixer wash cups, reagent carousel and water bath overflow waste area. It drains into the Low Concentration Waste Tank. 1 Left Side c4000 2 LCW Manifold 3 LCW Tank R Reagent Probes Wash Cups OF2 Water Bath Overflow OF1 Water Bath Overflow STR Mixer Wash Cups S Sample Probe Wash Cup RC Reagent Carousel WMU Low Concentration Waste (LCW) Tank The WMU contains a low concentration waste tank that collects low concentration waste from the LCW Drain Manifold, cuvette washer vacuum pump (1&2) and air bleed lines. The LCW Tank drain line is gravity fed to external waste. The tank contains a float sensor for verifying that the LCW tank is draining correctly. 1 Air Bleed 2 VP 1 & 2 3 LCW Manifold 4 LCW Tank 1 Float Sensor Degasser Unit Water is distributed from the WMU open water tank to the Degasser Unit. The Degasser Unit contains a degassing vacuum chamber (canister), vacuum pump, sensor and control board that is used to remove bubbles in the DI water for use by the Probe Wash Pump, Cuvette Wash Pump, Wash Solution Pump, and Water Bath Fill Pump. The Degasser unit is compatible with the ARCHITECT c8000 and c16000. 1 Heater Assembly 2 Water Filter 3 Degasser Unit 4 Water Bath Fill Pump Water Bath Fill and Water Heater Water Bath Fill Water from the Degasser Unit is pumped through the Heater Assembly, water filter, manifold and into the water bath by the peristaltic Water Bath Fill Pump. A float switch located in the water bath tells the c4000 module when the water bath is full. Water Bath Heater Once the water bath is full of water, the Water Bath Fill Pump turns off and the Water Bath Heater Assembly turns on. The Water Bath Heater Assembly contains a magnetic pump (MP), heater and radiator. When the unit is on, it maintains a constant water temperature of 37°C ± 0.2°C by re-circulating the water from the bath through a manifold, into the radiator, heater, and back to the water bath in a closed loop. The manifold is also used to provide re-circulated heated water to the ICT Warming Jacket to keep the ICT module warm. Sample & Reagent Lines - Internal Probe Wash Probe Wash Pump To wash the inside of the sample and reagent probes, the WMU supplies water through a manifold to the Degasser Unit that removes air bubbles from the DI water. On the down stroke of the internal probe wash pump, water flows from the degasser through a check valve and into the bellows style pump. On the upstroke, water is forced out the top of the pump through a check valve and into a valve unit that splits the water flow to the COM ports of the sample valve, reagent 1 valve and reagent 2 valve. When it is necessary to wash the probes internally, the sample and reagent valves allow water to pass through the NC ports to the sample and reagent syringes and out to the probes. As the pump cycles, water is flushed through the probes and into the probe wash cups. Upon completion of a wash cycle, the valves close the NC port (creating a slight pressure in the tubing between the Probe Wash Pump and the valves) and open the NO port. The line pressure forces water to the Water Bath Overflow Area. 1 Water Bath Overflow Area 2 From the NO Ports on Probe Wash Valve Unit Cuvette Washer Fluidics Overview An eight step cuvette washing process is used to clean the cuvettes on the c4000 module. The cuvette washer aspirates sample and reagent from previous reactions to a high concentrate waste container or drain. It washes the cuvettes with alkaline and acid wash solutions and provides two water rinses. After the water rinses, it dispenses water for a water blank reading and dries the cuvette prior to the next sample. 1 High Concentration Waste (HCW) Pump- evacuates HCW from the cuvette. 2 Wash Manifold- provides wash solutions to the nozzles. 3 LCW Manifold- used to evacuate wash waste (LCW) to Vacuum Pump 2. 4 Washer Nozzles- eight nozzles used to aspirate waste, wash and dry cuvettes Cuvette Washer Nozzles Cuvette Washer Nozzles Description Nozzle 1 Aspirates the high concentration waste through a high concentration waste peristaltic pump. The nozzle also dispenses DIwater into the cuvette from the cuvette wash bellows pump during the aspiration. Water is dispensed through small holes in the welded area in the mid-section of the nozzle. Nozzle 2 Delivers a 1.3% mixture of alkaline solution and DI water that is created and provided by the Wash Solution Pump using a micro syringe block and 1 mL syringes. After the alkaline solution is dispensed into the cuvette, it is aspirated through a vacuum pump. Nozzle 3 Delivers a 0.9% mixture of acid solution and DI water that is also created by the Wash Solution Pump. The alkaline and acid solutions are located on the weighted platform scales in the front of the analyzer. Nozzles 4/5 Dispense DI water into the cuvettes and aspirate the water through a vacuum pump. Nozzle 6 Has two nozzles, one providing DI water that fills the cuvette to a certain level and the other to aspirate part of the water out of the cuvette. The amount of water left in the cuvette is used by the system for a water blank which is measured as the cuvette passes through the optics unit. The transmission or intensity of the light measured in the water blank is used as the reference value for the subsequent measurement of the sample in this cuvette. It is also used as a cuvette integrity check. Nozzle 7 Performs an aspiration of the water blank. Nozzle 8 Has a porous cuvette drying tip installed on the nozzle that removes water from the cuvette prior to introduction of the next sample. The nozzle is spring loaded which allows the nozzle to move up when the drying tip is driven into the physical bottom of the cuvette. This ensures the cuvette drying tip is touching the bottom of the cuvette. Water in the bottom or on the side walls of the cuvette is aspirated by a vacuum pump through small holes on the side and bottom of the nozzle shaft. Cuvette Washing - Water Wash Cuvette Wash Pump The bellows style Cuvette Wash Pump supplies degassed water to the High Concentration Waste Nozzle, Water Wash Nozzle 1, Water Wash Nozzle 2, and the Water Blank Nozzle. Degassed DI water is used to prevent water from dropping out of the nozzle tip due to bubbles in the tubing. During the cuvette wash process, water from the Degasser Unit is directed through a check valve and into the bellows on the down stroke of the pump. On the up stroke of the pump, the cuvette wash valve opens the NC port allowing water to flow through a check valve and to a manifold that distributes water to the nozzles for cuvette rinsing. During normal running, cuvette washing is a continuous process. Upon completion of a wash cycle, the cuvette wash valve closes the NC port creating a slight pressure in the tubing between the Cuvette Wash Pump and the valve. The NO port opens and relieves any pressure or air in the line to the WMU open tank. Cuvette Washing - Alkaline and Acid Wash Solutions Wash Solution Pump As part of the cuvette wash process, alkaline and acid solutions are introduced into the cuvette aspirated. The Wash Solution Pump dilutes the alkaline solution to 1.3% and the acid solution to 0.9% and distributes the solutions to the nozzles for the cuvette washing. On the down stroke of the pump, degassed water from the Degasser Unit is supplied to the two 1mL syringes located on the front of the pump through check valves mounted at the top of each syringe. At the same time, the ALK1 and ACD1 valves turn on (NC is open, NO is closed) allowing aspiration of straight alkaline and acid solution from the bulk solution bottles into the micro syringes that are also mounted on the pump. The 1mL syringes filled with water and small, measured amounts of alkaline and acid solutions in the tubing lines create the proper dilutions to be delivered to the cuvette washer nozzles. On the up stroke of the pump, valves ALK1 and ACD 1 turn off (NC is closed, NO is open) and valves ACD2 and ALK2 turn on (NC is open, NO is closed), allowing the solutions to be pumped into the nozzles 2 and 3 of the cuvette washer. At the completion of the cycle, valves ALK2 and ACD2 turn off (NO open, NC closed), allowing any remaining solution or air to be directed to the low concentration waste tank through the valves normally open port. Cuvette Washing - Water & Wash Solution Aspiration and Drying Vacuum Pumps After the cuvettes have had alkaline, acid or water introduced during the wash process, they must be emptied and dried prior to the next sample dispense. To achieve this, a duel-headed vacuum pump is used to remove the contents of the cuvettes and provide a final drying before the cuvettes are used again. The vacuum pump has two heads that are identified as VP1 [1] and VP2 [2]. 1 VP1 2 VP2 VP1 is attached directly to the cuvette dryer tip and provides final drying of the cuvette before it is used again. Aspirated water is directed through the vacuum pump to the LCW tank in the WMU. VP2 [2] aspirates water rinse, alkaline wash solution and acid wash solution from the cuvettes into the nozzles through a manifold and directs it to the LCW tank in the WMU. Low Concentration Waste and Drain Lines Low concentration waste from the reagent supply center, wash cups, and the water bath overflow area is directed to the LCW tank through a drain pipe type waste manifold. (The water bath overflow area collects waste from the ICT reference solution cup, the sample and reagent valves, and the water bath overflow.) Low concentration waste from the cuvette washer and the air bleed lines from the wash solution pump is also directed to the low concentration waste tank. Waste from the tank is directed out of the back of the c4000 module. A float switch is located inside the LCW tank to alert the operator of an overflow situation. The tank also has a separate overflow port to an external fitting should it overflow. High Concentration Waste HCW From Cuvette Upon completion of a sample run, the patient sample and reagent are first aspirated from the cuvette through nozzle #1 (HCW Nozzle) on the cuvette washer. Cuvette washer nozzle #1 (HCW nozzle) is connected directly to the peristaltic style HCW pump. When the pump rotates, patient sample and reagent are aspirated from the cuvette into the nozzle and sent to the external high concentration waste container or drain. HCW From ICT HCW Cup ICT waste is drawn from the ICT HCW Cup through a check valve and into the #14 syringe on the ICT aspiration pump during the down stroke of the pump. On the subsequent up stroke of the pump, the waste is pushed through the other side of the check valve to the external high concentration waste container or drain. ICT Aspiration Pump Fluidics Located on the right side of the ICT Aspiration Pump, is the number 11 syringe, or Aspiration Syringe. It is used to aspirate sample or ICT Reference Solution into the ICT Module for measurement. • On the down stroke of the ICT Aspiration Pump, the aspiration syringe draws down and aspirates fluid into the ICT Probe, through the ICT Module, into the normally closed (NC) port of the Pinch Valve, out of the common port (COM) and into the aspiration syringe. • On the upstroke of the pump, fluid is pushed up through the common port of the Pinch Valve, out of the normally open (NO) port, and into the ICT High Concentration Waste (HCW) Cup. Located on the left side of the ICT Aspiration Pump, is the number 14 syringe. It is used to pump waste from the ICT HCW Cup to an External HCW Container or drain. • As the ICT Aspiration Pump moves down, the number 14 syringe draws the high concentration waste from the ICT HCW Cup down through a check valve and into the syringe. • On the upstroke of the pump, the waste is pushed out through a check valve and to the External HCW Container or drain. ICT Reference Solution Pump Fluidics Located on the right side of the ICT Reference Solution Pump, is the number 12 syringe. It is used to aspirate and dispense ICT Reference Solution into the ICT Reference Solution Cup. • On the down stroke of the ICT Reference Solution Pump, ICT Reference Solution is aspirated from the bulk solution bottle, through a check valve and into the syringe. • On the upstroke of the pump, the ICT Reference Solution is pushed up through a check valve, through the Warming Ring and into the upper port of the ICT Reference Solution Cup. On the left side of the ICT Reference Solution Pump, is the number 13 syringe. It is used to pump fluid from the ICT Reference Cup to the Water Bath Waste Overflow area. • On the down stroke of the ICT Reference Solution Pump, the number 13 syringe aspirates the ICT Reference Solution from the bottom port of the ICT Reference Solution Cup down through a check valve and into the syringe. • On the upstroke, it pushes the solution through the check valve and into the Water Bath Waste Overflow area. Key Diagrams, M&Ds and Alignment/Adjustment Procedures Diagram/M&D/Procedure Description Fluidics Diagram Diagram of system fluidics. M&D 1175 Wash Cup Test Procedure for measuring and adjusting water flow for sample, reagent and mixer wash cups. M&D 2155 Flush Bulk Solution Primes ICT reference solution, alkaline wash and acid wash solutions fluidic lines. M&D 2132 Flush Water Lines Flushes all water lines. P-668 Cuvette Washer Check Verifies cuvette washer operation. P-621 Cuvette Washer Alignment Cuvette washer alignment procedure. PCB Card Cage PCBs The Card Cage is located behind the right front door and access panel. 1 Card Cage 2 Sensor LED Board CPU The CPU contains the c4000 module firmware and provides communication with the SCC. A backup of the robotics information is stored in the CPU board flash memory. AC/DC Controller The AC/DC Controller board circuitry provides +12VDC valve control, 100VAC motor control and I/O for the associated sensors. It also provides 8-bit A/D conversion for temperature and voltage monitoring. Temperature monitoring: • Reagent Supply Center (2-10×C) • Water Bath (36.8-37.2×C) • Ambient (15-30×C) • Perform M&D 3525 Temperature Status to view readings Voltage monitoring: • • • • • • • +5VDC (card cage) +15VDC (card cage) -15VDC (card cage) +12VDC (valves) +11.45VDC (lamp) +24V.VDC (stepper motors) Perform M&D 5712 Voltages Test to view readings. SM AC/DC Controller The SM AC/DC Controller board provides I/O for the, reagent carousel lid sensor and controls the CRA (Continuous Reagent Access). As an option, the system has a CRA that provides a reagent auto loading function that will load and unload reagents through the RSH without disrupting assay processing. SMC Controller The Stepper Motor Controller (SMC) board circuitry provides 2-phase DC stepper motor control, 5-phase DC stepper motor control and I/O for the associated sensors. Pressure Monitoring Board The Pressure Monitor board has a digital processor that monitors the electrical output signals from the pressure monitoring sensors during the sample and reagent aspiration, dispense and wash process. DAQ The DAQ board consists of a 16 bit A/D converter circuit and a CPU of the 386EX series. The DAQ board serves to amplify and convert the optical and ICT analog signals to a digital format to be sent to the SCC. The DAQ board also controls the operation of the c4000 module and stores the robotics alignment information in SRAM memory. Sensor (MECH) LED Board The Sensor LED board contains light emitting diodes that indicate the status of the sensor signals input to the AC/DC Controller, Stepper Motor Controller and SM AC/DC Controller boards. The Sensor LED board is located on the inside of the card cage cover and is connected to the three controller boards by cables. Card Cage Backplane The Card Cage Backplane (upper and lower) distributes signals and voltages to the card cage circuit boards. Driver Box Assembly PCBs The Driver Box Assembly is located behind the right front door and access panel. 1 SMD 2 Mixer Driver 3 AC/DC Driver Stepper Motor Driver (SMD) Board The SMD board drives the 2-phase and 5-phase +24VDC stepper motors by control signals (speed, direction and power) from the Stepper Motor Controller (located in the card cage) and +24VDC from the Main Power Supply. AC/DC Driver The AC/DC Driver board drives the 100VAC motors and +12VDC magnetic valves by ON/OFF control signals from the AC/DC Controller (located in the card cage) and 100VAC and +12VDC from the Main Power Supply. STR (Mixer) Driver The Mixer Driver board drives the R1 and R2 piezoelectric mixers used for mixing sample and reagent in the cuvettes. Note: STR is an abbreviation for stirrer. Mixer Drive frequency: • Cuvette- 120Hz • Wash Cup Up position- 70Hz • Wash Cup Lower Position- 120Hz Additional PCBs Mounted Throughout The System Liquid Level Sense (LLS) Board The LLS board [1] is located above the Card Cage. 1 LLS Board The LLS board is used to detect contact with liquid surface by the sample probe, R1 probe and R2 probe. Fluid sense and air detection is based on capacitance for the sample and reagent probes. Thresholds for liquid detection are adjustable with potentiometers located on the LLS board. CNN (Distribution) Board The CNN Board is mounted on the rear right side of the c4000 module. 1 CNN Board The CNN board is a distribution/interface for signals between the card cage boards and the circuit boards and sensors It also generates voltages used for the PM sensors and isolated voltage for the LLS Board. TBC (Temperature Bath Controller) The TBC circuit board is located inside of the Heater Assembly in the center rear of the c4000 module. 1 TBC 2 Ambient Temperature Sensor The water bath temperature is monitored and controlled by the Temperature Bath Controller board. The TBC monitors the status of the water bath temperature sensor and drives a heater and pump to maintain the water bath temperature at 37° C. The ambient temperature sensor is also connected to the TBC. ICT Pre- Amp Board and IREF Level Detector Board The ICT Pre- Amp Board and IREF Level Detector Board are located at the front, upper right of the c4000 module. 1 Remove keypad. 2 Remove screw. 3 Slide up. The ICT Pre-Amp board receives and amplifies the signals from the ICT module and transmits the signals to the DAQ board for processing. 1 ICT Pre Amp 2 IREF Level Detect Board The IREF Level Detect board is connected to electrode sensors contained in the ICT Reference Solution Cup. It is used to verify that ICT Reference Solution is dispensed into the cup and that ICT Reference Solution is aspirated out of the cup by the ICT probe. It is mounted on the side of the ICT Pre-amp board located at the front, upper right of the c4000 module. There are two Light Emitting Diodes (LED) located on the board that indicate the status of the fluid sensors in the ICT Reference Solution Cup. Schematic Diagram Wiring Diagram Link 7A_9003b.pdf Key Diagrams, M&Ds and Alignment/Adjustment Procedures Diagram/M&D/Procedure Description Option M&D 3805 Pressure Monitor Test Tests the pressure monitor board and pressure monitor transducer for communications and consistency. N/A M&D 3525 Temperature Status Displays the current and average temperatures of the water bath, reagent supply center and the instrument ambient. N/A M&D 5712 Voltages Test Displays/tests the voltages of +5V, -15V, +15V, +12V, +11.5 and +24V. N/A Power Supply Overview The Main Power Supply and Transformer are used to generate the required AC and DC voltages for the c4000 and RSH modules. The Main Power Supply is located at the right-rear of the c4000module. 1 Main Power Supply 2 220V Input 3 Main Power Switch 4 Voltage output to c4000 and RSH 5 Transformer The Main Power Supply input receives the incoming 220VAC line voltage and routes it through a main power switch, circuit protector, line filter and out to the primary side of the Transformer. The primary side of the Transformer is tapped to allow selection of the corresponding incoming voltage: AC200V, AC208V, AC220V, AC230V, and AC240V. The secondary side of the Transformer provides 24VAC and 100VAC to the Main Power Supply for the eight built-in sub power supplies that generate required DC voltages for the c4000 and RSH modules. Main Power Supply Output Voltages Power Supply Voltage and Function Adjustable PS1 ±12VDC (2A) Card Cage Backplane +5VDC (15A) Card Cage Backplane/CPU/Logic +3.3VDC (8A) Card Cage Backplane/CPU Yes PS2 ±15VDC (1A) Analog Power Supply (DAQ) Yes PS3 +12VDC (8A) Valve Drive Power Supply Yes PS4 +11.45VDC (2A) Lamp Power Supply Yes PS5 +12VDC (27A) Peltiers and fan for Reagent Supply Center Yes Note ON/OFF controlled by thermostat. PS6 +24VDC (18A) c4000 Stepper Motor Drive Power Supply Yes PS7 +36VDC (18A) RSH Motor Drive Power Supply Yes PS8 +12VDC (2A) RSH Communications -12VDC (1A) RSH Communications +5VDC (10A) RSH CPU Control Yes 100VAC from Transformer AC Motors, Pumps and Heaters N/A 24VAC From Transformer Piezo Mixer Drive and Temperature Control Circuit N/A Main Power Supply Panel PS Power Supply LF Line Filter 220V Input TB Terminal Block NF Noise Filter MC Magnetic Contactor The panel of the Main Power Supply contains: 1. Circuit protectors/breakers: • • • • • AC2: TBC,STR-DRIVER AC24V Line AC3: AC/DC DRIVER AC100V Line AC4: TBC AC100V Line AC5: Cooling FAN AC100V Line CP1: SMD DC+24V Line1 2. • CP2: SMD DC+24V Line2 • CP3: SMD DC+24V Line3 • CP4: SMD DC+24V Line4 • CP5: SMD DC+24V Line5 • CP6: SMD DC+24V Line6 • CP7: SMD DC+24V Line7 Light Emitting Diodes (LEDs) for each power supply voltage (illuminated indicates on). 3. Variable resistors for adjusting voltage output. 4. Switches. Note The LED for PS5 is only illuminated when power is needed to cool the Reagent Supply Center (ON/OFF is controlled by the thermostat). Power Switches There are four power switches on the c4000 module: • Three Main Power Supply switches- Main Switch (Breaker) controls the line voltage into the Main Power Supply - ANA Switch- controls power to the c4000 analyzer - RSH Switch- controls power to the RSH • Maintenance power switch (left side of c4000 module) - removes power from the c4000 module (both standalone and integrated) except for the 12VDC used to cool the Reagent Supply Center and the RSH. Maintenance Switch Off c4000 and ci4100 Function Main Power Supply–Side Panel–Jumper Configuration Stops power to c4000 only. Provides power to RSH and c4000 Reagent Supply Center and allows maintenance be performed on the c4000 while the i1000 and RSH are processing samples. Note With the maintenance switch turned off, power supply LEDs are Off (not lit) for PS1 PS2 PS3 PS4 PS6 LEDs are On (lit) for: PS5 PS7 PS8 Main Power Supply Side Panel The side panel of the power supply provides connectors for AC and DC distribution to the c4000 and RSH modules. Connector Name Output/Function Transformer AC100V,AC24V AUTO Sub Power Supply ON/OFF Control AC100V Input AC100V AC100V Output CNBD CNN BOARD DC+5V,DC+12V Output ON/OFF Spare AC100V Output CPUM DC Voltage Monitor Output +24V1 SMD DC+24V Output(1) +24V2 SMD DC+24V Output(2) TBC TBC Unit DC+5V,AC24V Output R/F Front Power Control Switch +12V AC/DC Driver DC+12V Output STRD STR-Driver2 AC24V,DC±15V Output CPU DC+5V,DC±15V,DC+3.3V,DC±12V Output AC AC/DC DRIVER, Cooling FAN AC100V Output UC Unit Cooling DC+12V Output TBC Water Bath Control AC24V Output,+5V Output LAMP LAMP DC+11.5V Output MC IN Magnet Contact Input Side MC OUT Magnet Contact Output Side RSH RSH Power Supply Input Side Connector C+RSH c4000 Standalone RSH Connector i+RSH 4000 Integration RSH Connector J1 RSH DC+36V Output J2 RSH DC+36V Output J3 RSH DC+12V Output J4 RSH DC±12V Output J5 RSH DC+5V Output J6 RSH DC+5V Output Card Cage Backplane The Card Cage Backplane (upper and lower) distributes signals and voltages to the card cage circuit boards. Lower Backplane Terminal Blocks Upper Backplane Terminal Blocks The following voltages are supplied to the card cage circuit boards via the card cage back plane: • • • • • • • • +5VDC (For digital circuitry and voltage monitoring) +3.3VDC (For digital circuitry) ±12VDC (For digital circuitry) +15VDC (For analog circuitry and voltage monitoring) -15VDC (For analog circuit and voltage monitoring) +12VDC (For DC fans and Magnetic Valve voltage monitoring) +11.45VDC (For Lamp voltage monitoring) +24VDC (For SMD voltage monitoring) CNN (Distribution) Board The CNN board is a distribution/interface for signals between the card cage boards and the circuit boards and sensors mounted within the instrument frame. The CNN also generates the following voltages: • +5VDC: Voltage supply of various sensors • ±12VDC: Voltage supply of pressure sensors • ±15VDC: Voltage supply of LLS PCB A DC/DC converter generates the +5VDC and ±15VDC (isolated). The ±15V DC-DC converter for the LLS board is mounted on the CNN board providing separation from the DAQ ±15VDC. Therefore, the oscillation noise from the level sense board will not influence the DAQ board. Schematic Diagram Power Diagram Link 7A_9004.pdf Key Diagrams, M&Ds and Alignment/Adjustment Procedures Diagram/M&D/Procedure Description Total Wiring Diagram Provides a diagram of system wiring connection. Power Supply Diagrams Provides a diagram of the Main Power Supply. M&D 5712 Voltages Test Displays/tests the voltages of +5V, -15V, +15V, +12V, +11.5 and +24V at the AC/DC Controller (card cage). P-613 +5VDC (PS1) Measurement Perform this procedure to measure +5VDC (PS1) in the Main Power Supply. P-656 +3.3VDC (PS1) Measurement Perform this procedure to measure +3.3VDC (PS1) in the Main Power Supply. P-655 +12VDC and -12VDC (PS1) Measurement Perform this procedure to measure +VDC (PS1) in the Main Power Supply. P-645 DAQ +15/-15VDC (PS2 Power Supply) Measurement Perform this procedure to measure the +15/-15 VDC on the DAQ Board. P-614 +12VDC (PS3) Measurement Perform this procedure to measure +12VDC (PS3) in the Main Power Supply. P-616 +11.5 VDC, Lamp Power Supply (PS4) Measurement Perform this procedure to adjust DC Lamp Power Supply in the Main Power Supply. P-615 +12VDC (PS5) Measurement Perform this procedure to measure +12VDC (PS5) in the Main Power Supply. P-660 +24VDC (PS6 Power Supply) Measurement Perform this procedure to measure the +24V on the SMD. P-657 +36VDC (PS7 Power Supply) Measurement Perform this procedure to measure the +36V on the RSH. P-658 +5VDC (PS8 Power Supply) Measurement Perform this procedure to measure the +5V on the RSH. P-659 +12VDC (PS8 Power Supply) Measurement Perform this procedure to measure the +12V on the RSH. Valves, Motors and Sensors The c4000 module utilizes valves, motors and sensors to perform tasks associated with fluid delivery and movement of mechanical subassemblies. This is accomplished through the use of: • • • • • +12VDC Magnetic Valves 100VAC Motors +24VDC Stepper Motors (2-Phase and 5-Phase) Motor Assembly Position Sensors Robotic Alignments +12VDC Magnetic Valve Control The AC/DC Controller Board controls the +12VDC magnetic valves by sending ON/OFF control signals to the AC/DC Driver Board. The AC/DC Driver Board applies the +12VDC drive voltage to activate the magnetic valve. The following +12VDC magnetic valves are used on the c4000 module and can be manually controlled by performing M&D 5140 Pumps/Valves Test: • • • • • • • • • • • • Sampling Probe Valve Reagent 1 Probe Valve Reagent 2 Probe Valve DI Water Supply Valve Water Bath Supply Valve Mixer Wash Valve External Reagent Probe Wash Valve External Sample Probe Wash Valve Cuvette Dry Valve ICT Aspiration Pinch Valve ACD1/ALK1 Wash Valve ACD2/ALK2 Wash Valve 100VAC Motor Control and Sensors The AC/DC Controller Board controls the movement of the 100VAC motors by monitoring the motor assembly up/down sensor status and sending control data to the AC/DC Driver Board. The AC/DC Driver Board applies the 100VAC drive voltage to the motor. When an ON signal is sent by the controller board to the driver board, current is applied to the motor until an OFF signal is received. The CNN Board serves as an interface between the motor assembly sensor signals and the AC/DC Controller Board. The Sensor LED board, located in the card cage, contains light emitting diodes that indicate status of the sensor signals input to the AC/DC Controller. The following 100VAC motors are used on the c4000 module and can be manually controlled by performing M&D 5140 Pumps/Valves Test and M&D 1101 Robotics Test Tool: • • • • • • Magnetic Pump (WMU- external probe wash) Probe Wash Pump (internal probe wash) Cuvette Wash Pump Wash Solution Pump (alkaline and acid wash) ICT Reference Pump ICT Aspiration Pump • • • • Vacuum Pump (cuvette washer VP1 & VP2) Mixer Arm Up/Down Cuvette Washer Up/Down Reagent Carousel Cover Motor +24VDC Stepper Motor Control and Sensors The Stepper Motor Controller Board (SMC) controls the movement of the 2-phase and 5-phase stepper motors by monitoring the motor assembly home/pitch/position sensor status and sending direction, speed and power data to the Stepper Motor Driver Board (SMD). The SMD applies the +24VDC drive voltage to the 2-phase and 5-phase stepper motors. The CNN Board serves as an interface between the motor assembly sensor signals and the Stepper Motor Controller Board. The Sensor LED board, located in the card cage, contains light emitting diodes that indicate the status of the sensor signals input to the Stepper Motor Controller. The following +24VDC stepper motors are used on the c4000 module and can be manually controlled by performing M&D 1101 Robotics Test Tool. 2-Phase Stepper Motors • • • • • • • • • • • • • 5-Phase Stepper Motors R1 Arm Up/Down Motor R1 Arm Rotate Motor R2 Arm Up/Down Motor R2 Arm Rotate Motor R1 Mixer Rotate Motor R2 Mixer Rotate Motor ICT Arm Up/Down Motor ICT Arm Rotate Motor Sample Pump Motor R1 Pump Motor R2 Pump Motor HCW Peri Pump Motor Bath Fill Peri Pump Motor • Reaction Carousel Rotate Motor • Reagent Carousel Inner Motor • Reagent Carousel Outer Motor • Sample Arm Rotate Motor • Sample Arm Up/Down Motor The 5-phase stepper motors perform smaller movements than the 2phase stepper motors and are used on assemblies that require precise movement control. The following schematic provides greater detail for the overall motor operation: Schematic Diagram Valves, Motors and Sensors Diagram Link 7A_9005.pdf Motor Assembly Position Sensors Optical sensors and flags are used to determine the home, rotational, or vertical position of a motor driven device. A flag attached to the device interrupts the sensor signal by passing between the send and receive portions of the sensor. The interrupt signal indicates the presence of the flag within a specific position range of movement for that device. The sensor signal is reported to the CNN Board and monitored by a control circuit board contained in the card cage. Up, Down and Down OK optical sensors are used on AC motor driven devices. For most stepper motor driven devices, the sensors are labeled Home and Pitch. A flag interrupts both the Home and Pitch sensors when the DC motor assembly is in the home position. HOME SENSOR: Assembly starting or reference point. Flag must be in sensor at the same time Pitch Flag begins to enter the sensor. PITCH SENSOR: Tracks motor location. Flag must begin to enter the sensor at the same time Home Flag is in the sensor. The Sensor LED board, located in the card cage, contains light emitting diodes that indicate the status of the sensor signals from all DC motor and AC motor driven pump and robotic devices. Robotics Calibration The Stepper Motor Controller contains a factory set step table located in firmware. This table defines the default positioning step numbers for all robotic horizontal and vertical positions and is not user configured. These step values are generic and not instrument specific. The robotic calibration Mdse. train the robotic positioning by offsetting the SMC firmware generic step positions. This provides exact device positioning that is instrument specific. The offset position numbers are written into SRAM on the DAQ Board and backed up in flash memory on the CPU Board each time the user completes a robotic calibration M&D. There is no automated calibration of the sample pipettor at positions 1,2 and 3 in the sample solutions wash area. These positions are manually set utilizing M&D 1101 Robotic Test Tool (alignment data is saved to the DAQ and CPU when the user exits the M&D). The following is an example of the offset position step table that is shipped with a new c4000 system. M&D 6001 Save/Restore DAQ Data provides options to view, modify and print this information. R-Disk (Reaction Carousel) Offset Position Step Table -1 R-DISK -1 -2 -3 -4 -5 -6 -7 -8 -9 -0 Position Number: 1. Cuvette position in front of the light path. S-ARM UP 0 0 0 0 32 12 0 0 0 0 0 0 0 0 0 0 2 3 0 0 0 0 0 0 R1-ARM UP 0 0 0 0 -3 0 0 0 0 0 R1-ARM MV 0 0 0 0 0 0 0 0 R2-ARM UP 0 0 0 0 0 -5 -3 R2-ARM MV 0 0 0 1 0 1 4 STR1 0 0 STR2 0 0 R1C OUT -23 -4 3 0 0 R1C IN 28 3 0 0 0 ICTARM UP -16 -30 ICTARM MV 0 0 -1 -2 -3 -4 -5 -6 -7 -9 -0 S-ARM MV R1-ARM -8 MV=horizontal position UP=vertical position S-ARM (Sample Pipettor) R1-ARM (R1 Pipettor) R2-ARM (R2 Pipettor) Position Number: 1. [Not Used] Position Number: 1. [Not used] Position Number: 1. [Not used] 2. [Not Used] 2. [Not used] 2. [Not used] 3. Wash Cup 3. [Not used] 3. [Not used] 4. Resting Position 4. Cuvette 4. Cuvette 5. Cuvette Dilution 5. Wash Cup 5. Wash Cup 6. Cuvette Sample 6. Aspirate Reagent: Outer 6. Aspirate Reagent: Outer 7. [Not Used] 7. Aspirate Reagent: Inner 7. Aspirate Reagent: Inner 8. RSH 9. [Not Used] Each position has a horizontal (MV) and vertical (UP). Each position has a horizontal (MV) and vertical (UP). R1OC (Reagent Carousel) ICTARM (ICT Assembly) 10. Wash solution position 1 11. Wash solution position 2 12. Wash solution position 3 Each position has a horizontal (MV) and vertical (UP). STR (Mixer) Position Number: 1. Wash Cup Position Number: 1. Aspiration position Position Number: 1. ICT Reference Solution Cup 2. 2. Barcode reader position 2. 3. Bottle sensor position 4. Reagent loading position Cuvette Robotic Assembly Description Reagent Supply Center Cuvette The c4000 Reagent Supply Center contains two reagent carousels for the R1 and R2 reagents. The carousels are split into 15 segments and have a total of 90 positions for 55/20 mL size wedges, or 66 positions for 90 mL size wedges. 1 Inner carousel 2 Outer carousel 3 Fan The Reagent Supply Center provides 2-10 degrees C refrigerated storage for assay reagent packs utilizing four +12VDC thermoelectric devices (TEDs) located under the reagent carousel. 1 Fan 2 TED 3 TED A temperature sensor is used to detect the internal air temperature and is mounted into the bottom area of reagent carousel. A thermostat (ON/OFF control) is used to adjust the refrigerated temperature and a fan is used to exhaust the hot air from the TEDs. A 2D bar code reader is used to scan and identify the reagent type and position of each reagent cartridge loaded in the inner and/or outer carousel. The inner carousel and outer carousel drive units provide rotation for the reagent carousels. Each drive unit consists of a +24VDC five-phase stepper motor, gear and sensor assembly. 1 Inner Carousel Motor 2 Outer Carousel Motor 3 TED Reaction Carousel The c4000 reaction carousel contains 99 permanent, rectangular glass cuvettes in a temperature controlled water bath maintained at 37° C for a stable reaction process. The cuvettes are held in 9 cuvette holders (segments) attached to the reaction carousel ring. Each of the 9 segments contain 11 cuvettes. If necessary, all segments and individual cuvettes can be removed by the operator for cleaning or replacement. The reaction carousel rotates clockwise 1 ¼ turn every nine seconds during processing and positions the cuvettes for sample and reagent dispensing, mixing, photometry and cuvette washing. Each individual cuvette is read as it passes through the optical light path. The reaction carousel is gear driven by a +24VDC five-phase stepper motor and sensor assembly. Adjustable and fixed guide rollers provide proper tension and positioning of the carousel. 1 Adjustable and Fixed Guide Rollers 2 Motor Gear Cover 3 Motor Gear A water bath float sensor tells the c4000 module when the water bath is full. A cuvette wiper cleans the side of the cuvettes prior to them passing through the optical light path. The cuvette wipers are located at the rear of the water bath. The water bath has water input/output ports located on the underside. These ports provide water input, heated water exchange with the heater assembly and ICT module holder, and a drain. 1 VP 2 2 VP 1 The cuvette washer uses a 100VAC dual-headed vacuum pump to aspirate wash waste fluids and dry the cuvettes. Each pump head has a diaphragm mounted on a shaft to move liquid in and out of the pump. As the pump motor rotates, each diaphragm moves up and down, creating a vacuum in the fluidic lines, to aspirate fluid from the cuvettes. Each head has dual 0-rings and valves. Vacuum Pump 2 (VP-2) drains the low concentration waste liquid from the alkaline, acid, and DI water washes of the cuvettes (nozzles 2 - 7 of the cuvette washer). Vacuum Pump 1 (VP-1) provides vacuum for final drying of the cuvette, using nozzles 8. After drying, less than 2.0 ul residual water remains in the cuvette. Peristaltic Pumps 1 Head 2 Photo Sensor 3 Encoder 4 Motor 5 Pump Head Tubing There are two peristaltic pumps on the c4000: High Concentration Waste (HCW) and Water Bath Fill. Each pump is driven by a +24VDC 2-phase stepper motor and utilizes a photo sensor and encoder. The Water Bath Fill peristaltic pump supplies degassed water to the water bath. A check valve is installed between the pump and the reaction water bath to avoid leakage. The High Concentration Waste peristaltic pump removes high concentration waste from nozzle 1 on the cuvette washer. The pump is located near the wash station to prevent clogging by mixture of different reagents. The HCW and Water Bath Fill peristaltic pumps and pump heads are different; the components are not interchangeable. Cuvette Washer Assembly The upper section of the cuvette washer has eight nozzles, mounted on a platform, that are use in an eight-step cuvette washing process. The cuvette washer aspirates sample and reagent from previous reactions to a high concentrate waste container or drain. It washes the cuvettes with alkaline and acid wash solutions and provides two water rinses. After the water rinses, it dispenses water for a water blank reading, aspirates the water and dries the cuvette prior to the next sample. The lower mechanical section raises and lowers the nozzles into the cuvettes as the reaction carousel rotates. The lower mechanical section consists of a linear slide and 100VAC motor with three optical sensors to tell the c4000 module when the nozzles are in the correct up or down position. Mixer Assembly The mixer assembly consists of two main sections. A lower mechanical section, consisting of a 100VAC motor and linear slide, provides vertical drive to raise and lower the upper assembly. Three optical sensors in the lower section tell the c4000 module when the assembly is in the correct up or down position. The upper section consists of the mixers and two +24VDC stepper motors used to rotate the mixers horizontally over the reaction carousel. Optical sensors in the upper section provide feedback when the mixers are in the home position. The mixers are designed using piezoelectric effect. Piezoelectric effect is the characteristic of certain crystals to produce electricity when under strain or to produce vibration when a voltage at a specific frequency is applied. The mixers consist of a crystal attached to a thin mixer blade contained in a plastic enclosure. A voltage is applied to the mixers at two different frequencies depending on the desired mixer movement over the reaction carousel or wash cup. When in the cuvette or washing, a faster movement/shorter travel is required and the mixers are driven at 120Hz. When washing is complete, the mixers are driven at a lower frequency of 70Hz producing a slower movement/wider travel to remove water from the mixer blades. The Mixer Driver board drives the stirrers. The AC/DC and SM AC/DC controller and AC/DC Driver board provides control and drive to the vertical elevator. The SMC controller and SMD board control and drive rotation to the stirrers. The CNN board serves as the distribution board for monitoring various home sensor statuses. ICT Assembly The ICT assembly consists of a lower mechanical drive section and an upper section containing the ICT module. The lower mechanical section has two +24VDC stepper motors used to move the module in the horizontal and vertical directions along with optical sensors to tell the c4000 module when the unit is in the up/down or rotational home position. The upper section of the assembly contains the ICT module, holder and cover. The module holder has two ports that allow 37°C water to recirculate in a closed loop, indirectly stabilizing the temperature of the ICT module. Sample and Reagent Syringe Drives The syringe drives provide aspiration and dispense of sample and reagent. Each drive has a +24VDC two phase stepper motor. The motor rotates a ball screw mechanism to raise and lower the plunger located inside the attached syringe. Optical pitch and home (up) sensors are used to tell the c4000 module when the drive is in the home position. The sensors are also used to detect a single rotation of the motor. The amount of up and down travel the plunger moves determines the volume of sample or reagent dispensed or aspirated. Sample and Reagent Pipettors The sample and reagent pipettors are used to aspirate and dispense sample, reagent and solutions. Tubing from the probe and wires from the pre-amp board run down the center shaft of the pipettor, and into a coiled spring enclosure (to avoid damage during movement). The pipettors are belt driven in the horizontal and vertical direction by +24VDC stepper motors. The sample pipettor uses five phase motors and the reagent pipettors use two phase stepper motors. The motor and belt assembly uses optical home and pitch sensors to tell the c4000 module when the pipettor is in the home position. Pipettor positions for the reaction carousel, wash cups and reagent supply centers are software trained from the home position and rotation is limited by mechanical stops. All pipettors incorporate a crash or guard sensor to tell the c4000 module when an obstacle has been detected during downward travel. Bellows Pump There are two bellows style pumps used in the c4000 module. The two pumps are the internal probe wash and cuvette wash pump. The two pumps function in a similar manner but are not interchangeable. A 100VAC motor drives each pump and moves a bellows assembly up and down to provide aspiration and dispense of fluid. On the down stroke of the pump, fluid is drawn in on the side port and on the up stroke of the pump, fluid is directed out of the top port. This is accomplished by the use of directional poppet valves that permit fluid flow in only one direction. Optical sensors mounted inside the pump tell the c4000 module when the pump is in the up or down position. Wash Solution Pump The wash solution pump uses an AC motor, gear head assembly, cam and lifter to move the two 1mL syringes and the two micro syringes. The syringes create the 1.3% alkaline and 0.9% acid solutions used by the cuvette washer. The plunger of each syringe is attached to the lifter assembly that is driven by the cam located inside the lifter. As the cam rotates, the lifter and syringes are moved up and down to provide the proper dilutions. The micro syringes are contained in a syringe block and use the same tip seals and o-rings as the sample syringe. Optical sensors mounted inside the pump tell the c4000 module when the pump is in the up or down position. ICT Reference Solution (Iref) Pump The ICT Reference Solution Pump uses an AC motor, gear head assembly, cam and lifter to move the two attached 1mL syringes. These syringes are used to transport Iref solution to the ICT Reference Cup and remove the sample and Iref solution to waste. The plunger of each syringe is attached to the lifter assembly that is driven by the cam located inside a cut out slot in the middle of the lifter. As the cam rotates, the lifter and syringes are moved up and down. The Iref pump only has one sensor to detect full rotation of the motor. ICT Aspiration Pump The ICT aspiration pump uses an AC motor, gear head assembly, cam and lifter to move the two 1mL syringes. The syringes are used to aspirate sample or ICT Reference Solution into the ICT Module for measurement and to pump ICT high concentration to an external container or drain. The plunger of each syringe is attached to the lifter assembly that is driven by the cam located inside a cut out slot in the middle of the lifter. As the cam rotates, the lifter and syringes are moved up and down. The aspiration pump has a pinch valve attached to the frame of the pump. It also has two optical sensors to tell the c4000 module when the pump is in the up or down position. Key Diagrams, M&Ds and Alignment/Adjustment Procedures Diagram/M&D/Procedure Description Wiring Diagram Provides a diagram of the system wiring. M&D 5140 Pump/Valves Test Allows the user to turn on and off the c4000 pumps and valves. M&D 6001 Save/Restore DAQ Data 1. Save robotics alignment data from the DAQ board to the CPU board. 2. Load robotics alignment data from the CPU board to the DAQ board. 3. Clear all data from the DAQ board and load robotics alignment data from the CPU board to the DAQ board. 4. M&D 1101 Robotics Test Tool Review and enter alignment data Allows the user to manipulate and set offset positioning for the following: 1. Reaction carousel 2. R1 pipettor 3. R2 pipettor 4. Sample pipettor 5. Mixer 6. Reagent supply center 7. ICT 8. Wash station 9. Water bath exchange P-618 ICT Alignment Aligns ICT positions using alignment tools #1 and #2. P-620 Mixer Alignment Aligns the mixer positions using the 7mm mixer height tool. P-621 Cuvette Washer Alignment Aligns the cuvette washer positions using alignment tool #3. P-622 Sample Pipettor Head Horizontal Alignment Perform this procedure for horizontal alignment (manually) of the sample pipettor head. P-608 Sample Wash Cup Position Alignment Perform this procedure to align (mechanically) the Sample Wash Cup to Sample Probe. 1120 Sample Pipettor Calibration This procedure determines the position of calibration targets in relation to the home position of the sample pipettor P-623 Reagent Pipettor Head Horizontal Alignment Perform this procedure for horizontal alignment (manually) of the reagent pipettor heads. P-610 Reagent Wash Cup (R1 and R2) Alignment Perform this procedure to align (mechanically) the Reagent Wash Cups to Reagent Probes. 1121 R1 Pipettor Calibration This procedure determines the position of calibration targets in relation to the home position of the pipettor. 1122 R2 Pipettor Calibration This procedure determines the position of calibration targets in relation to the home position of the pipettor. P-611 Mixer (1 and 2) Wash Cup Position Alignment Perform this procedure to align (mechanically) the Mixer Wash Cups to the mixers. Liquid Level Sense (LLS) The liquid level sense system is capacitance based where the sample or reagent acts as a capacitor in series with the sample or reagent probe. A 32 KHz sine wave signal is applied to the probe and is also used as a reference signal input to a multiplier chip located on the LLS PCB. The 32 KHz sine wave signal from the probe is multiplied with the 32 KHz sine wave from the sine wave generator (reference signal). If the probe does not encounter liquid, there will not be a significant phase shift on the 32 KHz sine wave signal from the probe. When two sine waves that are in phase are multiplied, the output voltage is approximately zero, signifying that liquid has not been detected. If the probe encounters liquid, there will be a phase shift in the 32 KHz sine wave signal from the probe because the liquid and base plate of the instrument has some capacitance which causes a phase shift in the signal. When two sine waves that are out of phase are multiplied, an output voltage is generated signifying that liquid has been detected. LLS Board The LLS board is used to detect contact with liquid surface by the sample probe, R1 probe and R2 probe. The LLS board is located above the Card Cage. The thresholds for liquid detection are adjustable with potentiometers located on the LLS board. The CNN PCB generates the +5V and ±15V (isolated) used by the LLS PCB. This provides separation from the DAQ ±15V. This ensures that the oscillation noise from the level sense board does not influence the DAQ board. The threshold voltages for the Sample Pipettor and Reagent Pipettor are adjustable on the LLS PCB located above the Card Cage. Sample Pipettor LLS Errors The Sample Pipettor LLS circuitry is monitored as the Sample Pipettor steps down into the tube/cup. During this travel and when in fluid, the LLS signal is monitored for any change. If liquid is detected, the Sample Pipettor will step down an additional 3mm prior to aspiration. • “Liquid Not Found” error is generated when the Sample Pipettor travels to the lower step limit without detecting liquid. • “Liquid Contact Broken For Sample Pipettor” error is generated when (after liquid detection and aspiration into the probe) the LLS circuitry detects a change of phase, due to the loss of liquid (prior to dispense). • “Sample Pipettor Error While Moving Down” error is generated when the Crash Sensor is activated as the Sample Pipettor steps down to detect liquid. • “Liquid Too High For Sample Pipettor” error is generated when the LLS circuitry detects fluid at a step position that is higher than the top of the cup. • “Liquid Too Low For Sample Pipettor” error is generated when air is detected during sample aspiration. Reagent Pipettor LLS Errors The reagent pipettor LLS circuitry is monitored as the Reagent Pipettor steps down into a cartridge. During this travel and when in fluid, the LLS signal is monitored for any change. If liquid is detected, the Reagent Pipettor will step down an additional 5mm prior to aspiration. • “Liquid Not Found” error is generated when the Reagent Pipettor travels to the lower step limit without detecting liquid • ”Liquid Contact Broken For Reagent Pipettor” error is generated when (after liquid detection and aspiration into the probe) the LLS circuitry detects a change of phase, due to the loss of liquid (prior to dispense) • “Reagent Pipettor Error While Moving Down” error is generated when the Crash Sensor is activated as the Reagent Pipettor steps down to detect liquid • “Liquid Too High For Reagent Pipettor” error is generated when the LLS circuitry detects fluid at a step position that is higher than the top of the cartridge • Liquid Too Low For Reagent Pipettor” error is generated when air is detected during sample aspiration c4000 Pressure Monitoring Hardware Description A piezoresistive pressure transducer (Pressure Sensor) is situated in series with the sample and reagent probe tubing. The sensor is located on the Preamp Sensor PCB Assembly that contains a power supply and amplifier circuitry. The transducer generates an analog voltage that is proportional to the pressure change monitored. Analog electrical signals flow from the sensor to the PM Board located in the Card Cage. Analog signals are converted into 16 bit digital signals by an A/D converter located on the PM Board. (Digital electrical signals also flow from the LLS Board to the PM Board.) A digital signal processor (DSP) contains firmware that determines if the pressure signal from the transducer is within the specifications for the task (aspiration, dispense or probe rinse) being performed. The PM Board sends and receives data from the c4000 CPU via the Compact PCI bus. c4000 Pressure Monitoring Functional Description The c4000 Pressure Monitoring (PM) subsystem is operational during the aspiration, dispense and wash aspirations of the c4000 sample and reagent pipettor arm operations. The PM subsystem is designed to detect abnormalities such as clots (probe obstructions) and air bubbles within the patient sample, and air bubbles and large amounts of foam in reagent bottles during aspiration and dispense operations. It also monitors the wash pressure during the internal probe wash to determine if the wash pressure is below or above pressure thresholds referenced in the PM Board firmware. PM During Aspiration During sample and reagent aspiration, the pressure monitoring and liquid level sense circuitry is active as the probe descends toward liquid. If liquid is not sensed correctly, an LLS error–such as liquid not found, or liquid too high–is generated and pressure monitoring ceases for the aspiration event. Note An LLS error generates an accompanying PM error because the LLS error halts processing for the aspiration event, and results in an abnormal pressure waveform for the aspiration. If liquid is detected correctly, the pressure monitoring continues and the resultant PM waveform is analyzed for abnormalities. If a PM error is detected, the PM Board sends an error message to the c4000 CPU and SCC and aborts the processing of that sample. If a PM error is not generated, the c4000 continues to process the sample. PM During Dispense The PM subsystem is also active during the sample and reagent dispensing operations. If a PM error is detected during sample or reagent dispensing, the PM Board sends an error message to the c4000 CPU and SCC and aborts the processing of that sample. PM During Internal Probe Wash The PM subsystem monitors the pressure during reagent and sample internal probe wash operations to determine if excessively high or abnormally low pressures exist. Some of the abnormalities detected include clogged probes that create excessively high pressure, or major leaks in the tubing or failed wash pumps that can create low-pressure conditions. The c4000 performs wash operations before each run initiates. If a failure occurs during the wash sequence prior to the initiation of a run, the system stops, and the run terminates and generates a wash error. If a sample probe or reagent probe wash error occurs during a run, and cannot be resolved after 5 consecutive washes, the run terminates. Key Diagrams, M&Ds and Alignment/Adjustment Procedures Diagram/M&D/Procedure Description M&D 3805 Pressure Monitor Test This procedure will test the pressure monitor board and pressure monitor transducer for communications and consistency. 1127 Crash Sensor Alignment This procedure allows the user to align the crash sensor for either the sample probe, the R1 probe, or the R2 probe. P-617 LLS Voltage Adjustment Perform this procedure to adjust LLS threshold voltages on the LLS board. Temperature Control System The c4000 analyzer utilizes temperature sensors to monitor: • Internal ambient temperature • Water Bath temperature • Reagent Supply Center temperature Temperature readings from the Water Bath, Reagent Supply Center and Ambient sensor are monitored in real time and stored on the c4000 module until a request command is received from the SCC. Temperature data is used to calculate a moving average that is compared against defined normal ranges; an error is generated if specifications are exceeded. AC/DC Controller The AC/DC Controller PCB, located in the card cage, monitors voltage from temperature sensors, converts the voltage to a digital equivalent (A/D) and sends the digital data to the c4000 analyzer CPU. Temperature data is used to initiate heating or cooling and to generate errors if any of the three temperatures exceed specification. The SCC uses data to display temperatures in a readable format. The AC/DC Controller Board contains potentiometers on the top, front edge to adjust displayed temperatures on the SCC monitor. The potentiometer can be used to adjust the displayed temperature to match temperature output of a calibrated thermometer when needed. However, adjustment of the potentiometer does not affect the actual system temperature, it only changes the displayed temperature. Displayed Temp Adjustment Area Monitored Detector Location Reagent Supply Center Bottom area of Reagent Carousel 0.5 seconds Data gathered once per minute. Calculates moving average using 60 data points. 2-10°C AC/DC Controller VR1 Thermostat Water Bath Bottom of Water Bath 0.5 seconds Data gathered every 2 seconds. Calculates moving average using 5 data points. 36.8-37.2°C AC/DC Controller VR5 Heater Ambient Attached to Water Bath Heater 0.5 seconds Data gathered once per minute. Calculates moving average using 5 data points. 15-30°C AC/DC Controller VR4 N/A Reagent Supply Center Read Cycle Moving Average Calculation Temp Range Actual Temp Adjustment • • • • • The Reagent Supply Center provides 2-10°C refrigerated storage for assay reagent packs The Reagent Supply Center is cooled utilizing four 12VDC thermoelectric devices (TEDs) located under the reagent carousel A temperature sensor detects internal air temperature. It is mounted into the bottom area of reagent carousel A thermostat (ON/OFF control) is used to adjust the refrigerated temperature A fan exhausts hot air from TEDs Water Bath Heater Assembly The function of the Water Bath Heater Assembly is to circulate and heat degassed DI water in the Water Bath to 37°C +/- 0.2°C. Major components include: • • • • • • Radiator and fan Heater Water Bath Circulation Pump Temperature Bath Controller PCB (TBC) Ambient Temperature Sensor (thermistor) Water Bath Temperature Sensor (thermistor) Water Bath Circulation Pump Water circulation is provided by the Water Bath Circulation Pump (magnetic pump) and is constant as long as the water bath is full. If the Water Bath is NOT full, the Water Bath Circulation Pump and Heater shuts off. Water Bath Temperature Sensor Water temperature is monitored with the Water Bath Temperature Sensor located in the bottom of the Water Bath. Temperature Bath Controller (TBC) and Heater Water temperature is controlled by the TBC circuit board. 100VAC supplied to the TBC drives a heater element located inside a foam-jacketed water bottle. The element is fused and has a control thermistor installed for feedback. The fuse blows at 70°C. Water bath temperature is adjusted by rotating VR1 on the Temperature Bath Controller circuit board. Rotating clockwise increases temperature. Counter clockwise, lowers temperature. A jumper is included on the TBC circuit board for 30°C or 37°C water bath operation (c4000 uses 37°C only). Two LEDs on the TBC indicate the status of the heater: • CPU CMD (LED 2) displays the on/off signal status from the card cage • HEATER (LED 1) indicates when the heater is on/off and when temperature controls are stabilized. The HEATER (LED1) dims when stabilized or off. Radiator, Fan and Ambient Temperature Sensor The heater assembly has a fan/radiator section. The fan for the radiator turns on when water bath temperature reaches 37.2°C and ambient temperature reaches 28° C. An error is generated if water temperature is outside of the 36.7°C to 37.3°C range. An error is generated if ambient temperature is below 15°C or above 30°C. Both Fan and Heater are turned OFF under the following conditions: • • • • Water Bath is draining Water Bath is filling Water Bath is priming (air purge) System is booting up Key Diagrams, M&Ds and Alignment/Adjustment Procedures Diagram/M&D/Procedure Description M&D 3805 Pressure Monitor Test This procedure will test the pressure monitor board and pressure monitor transducer for communications and consistency. 1127 Crash Sensor Alignment This procedure allows the user to align the crash sensor for either the sample probe, the R1 probe, or the R2 probe. P-617 LLS Voltage Adjustment Perform this procedure to adjust LLS threshold voltages on the LLS board. ARCHITECT c4000 Service and Support Manual (Version 204733-123) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Troubleshooting (Document Control Number 204737-110) ARCHITECT c4000 Service and Support Manual (Version 204733-122) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Cable Routing and Connector Pin-out Troubleshooting Links Cuvette Wash Station Motor, Pump and Sensors Detergent Pump, Sensors and Valves ISE Motors and Sensors Reaction Carousel Motors, Sensors and Water Bath Valves Reagent 1 Arm Motors and Sensors Reagent 1 Syringe Pump Motors, Sensors and Valves Reagent 2 Arm Motor and Sensor Reagent 2 Syringe Pump Motors, Sensors and Valves Reagent Carousel Motors and Sensors Sample Arm Motors and Sensors Sample Syringe Pump Motor and Sensors Stirrer Motors and Sensors ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Cuvette Wash Station Motor, Pump and Sensors Cuvette Wash Station Motor and Sensor Cable Trace Description Code Card Cage Board Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable Cell Wash Station Up/Down Motor and Sensors Cell Wash CW Signal WS UP AC/DC CNN1-1 CNN34-1 CNN32-1 AD1 AC/DC CNN1-1 AC/DC CNN2-1 – CNN2-3 AD2,3 Cell Wash CCW Signal WS DWN AC/DC CNN1-2 CNN34-2 CNN32-2 AD1 AC/DC CNN1-2 AC/DC CNN2-1 – CNN2-3 AD2,3 Cell Wash WS BK Brake Signal AC/DC CNN1-3 CNN34-3 CNN32-3 AD1 AC/DC CNN1-3 AC/DC CNN2-1 – CNN2-3 AD2,3 Wash Station WSUP AC/DC CNN 7-1 CNN 2-1 CNN 18-24 CNN18-22 +5V CNN18-23 GND Wash Station WSDOWN AC/DC CNN 7-2 CNN 2-2 CNN 18-27 CNN18-25 +5V CNN18-26 GND Wash Station WSDOWNOK AC/DC CNN 7-3 CNN 2-3 CNN 18-30 CNN18-28 +5V CNN18-29 GND Cell Wash Pump and Sensors Cell Wash Pump CW Signal WCP ON AC/DC CNN1-18 CNN3418 CNN32-14 AD1 AC/DC CNN1-14 AC/DC CNN3-4 – CNN3-6 AD2,3 Cell Wash Pump Brake Signal WCP BK AC/DC CNN1-19 CNN3419 CNN32-15 AD1 AC/DC CNN1-15 AC/DC CNN3-4 – CNN3-6 AD2,3 Cuvette Wash Pump CWPUP AC/DC CNN 7-14 CNN 2-14 CNN 14-18 CNN14-16 +5V CNN14-17 GND Cuvette Wash Pump CWPDOWN AC/DC CNN 7-15 CNN 2-15 CNN 14-21 CNN14-19 +5V CNN14-20 GND AC/DC CNN9-1 CNN9-2 AD9 AC/DC CNN6-1 CNN6-2 AD6,7 Cell Wash Valve Cell Wash Valve CWV AC/DC CNN1-38 CNN3438 CNN32-34 AD1 AC/DC CNN1-34 Cell Wash Valve FB Signal CWV FB (SENS11) SM AC/DC CNN 8-47 CNN4-47 CNN32-58 AD1 AC/DC CNN1-58 CNN32-23 AD1 AC/DC CNN1-23 Cell Wash Drying (Vacuum Pumps) Cell Vacuum VP2 Pump ON Signal AC/DC CNN1-26 CNN3426 Vacuum SM CNN8-61 CNN 4-62 CNN 32-47 SENS24 (VP2 Pump Feedback FB) AC/DC Cell Dryer Pump ON Signal VP1 AC/DC CNN1-27 CNN3427 Dryer Pump Feedback SENS 25 (VP1 FB) SM AC/DC CNN8-62 CNN 4-50 CNN 32-48 CNN32-24 AD1 AC/DC CNN1-24 AC/DC CNN7-1 CNN7-4 AD6,7 High Concentration Waste Pump Motor and Sensors HCW Pump Motor NA PPHCINA SMC CNN3-36 CNN3636 CNN33-66 SM1 CNN1-66 CNN12-7 CNN12-12 SM11,12 HCW Pump Motor NB PPHCINB SMC CNN3-37 CNN3637 CNN33-67 SM1 CNN1-67 CNN12-7 CNN12-12 SM11,12 HCW Pump Motor TDA PPHCTDA SMC CNN3-38 CNN3638 CNN33-68 SM1 CNN1-68 CNN12-7 CNN12-12 SM11,12 HCW Pump Motor TDB PPHCTDB SMC CNN3-39 CNN3639 CNN33-69 SM1 CNN1-69 CNN12-7 CNN12-12 SM11,12 HCW Pump Motor Hold PPHCHOLD SMC CNN3-40 CNN3640 CNN33-70 SM1 CNN1-70 CNN12-7 CNN12-12 SM11,12 HCWP ISOHOME SMC CNN9-28 CNN 3-28 CNN 28-3 CNN28-1 +5V CNN28-2 GND HCWP ISOPITCH SMC CNN 9-29 CNN 3-29 CNN 28-6 CNN28-4 +5V CNN28-5 GND HCWP ISOPOS SMC CNN 9-30 CNN 3-30 Schematic Diagram Cuvette Wash Station Motor, Pump, and Sensors Link 7A_9017a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Detergent Pump, Sensors and Valves C4000 Detergent Pump, Sensors and Valves Description Code Card Cage Board Card Cage Backplane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC.DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd I/F Signal Cable Pins Driver Box Cable Detergent Pump Detergent Pump CW Signal DTGPON AC/DC CNN 1-16 CNN 34-16 CNN 32-20 AD 1 CNN1-20 CNN 5-1 – CNN 5-3 AD 4,5 Detergent Pump Brake Signal DTGBK AC/DC CNN 1-17 CNN 34-17 CNN 32-21 AD 1 CNN 1-21 CNN 5-1 – 5-3 AD 4,5 Detergent Pump Sensor DPUP AC/DC CNN 7-12 CNN 2-12 CNN 14-1 +5V CNN 14-2 GND CNN14-3 Detergent Pump Sensor DPBK AC/DC CNN 7-13 CNN 2-13 CNN 14-4 +5V CNN 14-5 GND CNN 14-6 Acid Detergent Sensor ESB3 AC/DC CNN 7-72 CNN 2-72 CNN 14-10 +5V CNN 14-11 GND CNN 14-12 Alkaline Detergent Sensor ESB14 AC/DC CNN 7-71 CNN 2-71 CNN 14-13 +5V CNN 14-14 GND CNN14-15 Alkaline Detergent Sensor Status ESB14SS AC/DC CNN 7-74 Acid Detergent Sensor Status ESB3SS AC/DC CNN 7-75 Acid & Alkaline Detergent Valve 1 ON Signal DTG1V AC/DC CNN 1-39 CNN 34-39 CNN 32-32 AD 1 CNN 1-32 CNN 8-9 CNN 8-10 Acid & Alkaline Detergent Valve 1 ON FB Signal DTG1V FB SM AC/DC (SENS 12) CNN 8-48 CNN 4-48 CNN 32-56 AD 1 CNN 1-56 Acid & Alkaline Detergent Valve 2 ON DTG2V CNN 1-40 CNN 34-40 CNN 32-33 AD 1 CNN 1-33 AC/DC CNN 8-11 CNN 8-12 AD 8 Signal Acid & Alkaline Detergent Valve 2 ON Signal DTG2VFB (SENS 13) SM AC/DC CNN 8-49 Schematic Diagram Detergent Pump, Sensors, and Valves CNN 4-49 CNN 32-57 AD 1 CNN 1-57 Link 7A_9018a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. ISE Motors and Sensors ISE Motor and Sensor Cable Trace Description Code Card Cage Board Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable ISE Arm Up/Down Motor and Sensors ISE Arm Up/Down Motor NA IAZINA SMC CNN2-6 CNN35-6 CNN33-41 SM1 SMD CNN1-41 SMD CNN10-1 – CNN10-6 SM10,11 ISE Arm Up/Down Motor NB IAZINB SMC CNN2-7 CNN35-7 CNN33-42 SM1 SMD CNN1-42 SMD CNN10-1 – CNN10-6 SM10,11 ISE Arm Up/Down Motor TDA IAZTDA SMC CNN2-8 CNN35-8 CNN33-43 SM1 SMD CNN1-43 SMD CNN10-1 – CNN10-6 SM10,11 ISE Arm Up/Down Motor TDB IAZTDB SMC CNN2-9 CNN35-9 CNN33-44 SM1 SMD CNN1-44 SMD CNN10-1 – CNN10-6 SM10,11 ISE Arm Up/Down Motor Hold IAZHOLD SMC CNN2-10 CNN35-10 CNN33-45 SM1 SMD CNN1-45 SMD CNN10-1 – CNN10-6 SM10,11 ISE Arm U/D IAEHOME SMC CNN9-2 CNN 3-2 CNN 27-6 CNN27-4 +5V CNN27-5 GND ISE Arm U/D IAEMIR SMC CNN 9-3 CNN 3-3 CNN 27-9 CNN27-7 +5V CNN27-8 GND ISE Arm Rotation Motor and Sensors ISE Arm IAHINA Rotate Motor NA SMC CNN2-1 CNN35-1 CNN33-46 SM1 SMD CNN1-46 SMD CNN10-7 – CNN10-12 SM10,11 ISE Arm IAHINB Rotate Motor NB SMC CNN2-2 CNN35-2 CNN33-47 SM1 SMD CNN1-47 SMD CNN10-7 – CNN10-12 SM10,11 ISE Arm IAHTDA Rotate Motor TDA SMC CNN2-3 CNN35-3 CNN33-48 SM1 SMD CNN1-48 SMD CNN10-7 – CNN10-12 SM10,11 ISE Arm IAHTDB Rotate Motor TDB SMC CNN2-4 CNN35-4 CNN33-49 SM1 SMD CNN1-49 SMD CNN10-7 – CNN10-12 SM10,11 ISE Arm IAHHOLD Rotate Motor Hold SMC CNN2-5 CNN35-5 CNN33-50 SM1 SMD CNN1-50 SMD CNN10-7 – CNN10-12 SM10,11 ISE Arm Rotation SMC CNN 9-1 CNN 3-1 CNN 27-3 CNN27-1 +5V CNN27-2 GND IAAHOME ISE Suction (Aspiration) Pump Motor and Sensors ISE Suction Pump CW Signal ICTP ON AC/DC CNN1-20 CNN34-20 CNN32-16 AD1 AC/DC CNN1-16 AC/DC CNN4-1 – CNN4-3 AD4,5 ISE Suction ICTP BK AC/DC CNN1-21 CNN34-21 CNN32-17 AD1 AC/DC AC/DC AD4,5 Pump Brake Signal ISE Aspir Pump ISPUP AC/DC CNN7-16 CNN2-16 CNN25-23 CNN25-21+5V CNN25-22 GND ISE Aspir Pump ISPDOWN AC/DC CNN7-17 CNN2-17 CNN25-26 CNN25-24 +5V CNN25-25 GND CNN1-17 CNN4-1 – CNN4-3 ISE Reference Pump Motor and Sensors ISE Ref Pump CW Signal ICTR ON AC/DC CNN1-22 CNN34-22 CNN32-18 AD1 AC/DC CNN1-18 AC/DC CNN4-4 – CNN4-6 AD4,5 ISE Ref Pump CCW Signal ICTR BK AC/DC CNN1-23 CNN34-23 CNN32-19 AD1 AC/DC CNN1-19 AC/DC CNN4-4 – CNN4-6 AD4,5 ISE Ref Pump IMPUP AC/DC CNN7-18 CNN2-18 CNN25-29 CNN25-27 +5V CNN25-28 GND AC/DC CNN8-7 CNN8-8 AD8 AC/DC CNN10-1 CNN10-2 AD10 ISE Aspiration Pinch Valve and Feedback ISE Suction Pinch Valve ON Signal IAVP AC/DC CNN1-32 CNN34-32 CNN32-31 AD1 AC/DC CNN1-31 ISE Suction Pinch Valve ON Signal FB SENS5 (IAV FB) SM AC/DC CNN8-41 CNN4-41 CNN32-55 AD1 AC/DC CNN1-55 ICT-Level Sensor ON Signal IREF AC/DC CNN1-49 CNN34-49 CNN32-40 AD1 AC/DC CNN1-40 ICT-Level (Ref Cup) Sensor FB Signal IREF FB SM AC/DC CNN8-58 CNN4-58 CNN32-64 AD1 AC/DC CNN1-64 ISE Ref Solution Sensor Status ESMCCSS IRECUP AC/DC CNN7-73 CNN3-73 CNN28-7 +5V CNN28-8 GND CNN28-9 ISE Ref Solution Sensor ESMCC AC/DC CNN7-70 CNN2-70 CNN14-9 CNN14-7 +5V CNN14-8 GND Schematic Diagram ISE Motors and Sensors Link 7A_9019a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Reaction Carousel Motors, Sensors and Water Bath Valves Reaction Carousel Motor and Sensor Cable Trace Description Code Card Cage Card Cage Back CNN Board Board Plane Cable Cable CNN Board Cable Driver Box Cable SMD or SMD or AC/DC or AC/DC or Mixer Bd Mixer Bd I/F Signal Cable Pins Cable Pins Driver Box Cable Reaction Carousel Motor and Sensors Reaction Carousel Rotate Motor CW RCCW SMC CNN3-51 CNN36-51 CNN33-1 SM1 SMD CNN1-1 SMD CNN2-1 – CNN2-5 SM2,3 Reaction Carousel Rotate Motor CCW RCCCW SMC CNN3-52 CNN36-52 CNN33-2 SM1 SMD CNN1-2 SMD CNN2-1 – CNN2-5 SM2,3 Reaction Crsl RUHOME SMC CNN 9-33 CNN 3-33 CNN 18-34 +5V CNN 18-35 GND CNN18-36 Reaction Crsl RUMIR SMC CNN 9-34 CNN 3-34 CNN 18-31 +5V CNN 18-32 GND CNN 18-33 Water Bath Heater ON/OFF IHTRON AC/DC CNN 7-77 CNN2-77 CNN13-3 Water Bath Cycle Pump ON/OFF IPMON AC/DC CNN 7-76 CNN2-76 CNN13-4 Water Bath Fan Cntrl ON/OFF IFANON AC/DC CNN 7-80 CNN2-80 CNN13-6 Water Bath Mtr Vlv Open IMVOPEN AC/DC CNN 7-78 CNN2-78 CNN13-7 Water Bath Mtr Vlv Clsd IMVSHUT AC/DC CNN 7-79 CNN2-79 CNN13-8 Water Bath Fan Ctrl Thrm IVRSTR+ AC/DC CNN 7-89 CNN2-89 CNN13-9 Water Bath Fan Ctrl Thrm IVRSTR- AC/DC CNN 7-90 CNN2-90 CNN13-10 Water Bath Full BATHLVL Detect AC/DC CNN 7-58 CNN3-58 CNN18-38 Water Bath Temp Sensor BATHTMP- AC/DC CNN 7-84 CNN3-84 CNN18-39 Water Bath Temp Sensor BATHTMP+ AC/DC CNN 7-83 CNN3-83 CNN18-40 DIW Supply Pump 1 (Water Bath) TWMP AC/DC CNN1-28 CNN9-11 CNN9-12 AD 9 Water Bath DIW Supply Pump 1 (Water Bath) FB SENS 26 SM AC/DC CNN8-63 DIW Valve 4 (Water Bath Supply) WBSV AC/DC CNN1-36 CNN34-36 CNN32-38 AD1 AC/DC CNN1-38 DIW Valve 4 Feedback WBSV FB (SENS 9) SM AC/DC CNN8-45 CNN4-45 CNN32-62 AD1 AC/DC CNN1-62 DIW Valve 5 (Pressurized DIW Control) PWV AC/DC CNN1-37 CNN34-37 CNN32-39 AD1 AC/DC CNN1-39 DIW Valve 5 Feedback SENS 10 SM AC/DC CNN8-46 CNN4-46 CNN32-63 AD1 AC/DC CNN1-63 Schematic Diagrams Reaction Carousel Motors, Sensors and Water Bath Valves CNN9-9 CNN9-10 AD 9 CNN9-9 CNN9-10 AD 9 Link 7A_9020a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Reagent 1 Arm Motors and Sensors Reagent 1 Arm Motor and Sensor Cable Tree Description Code Card Cage Board Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Driver Box Mixer Bd Cable Cable Pins R1 Arm Up/Down Motor and Sensors: R1 Arm Up/Down Motor NA R1ZINA SMC CNN2-66 CNN3566 CNN33-11 SM1 SMD CNN1-11 SMD CNN7-1 – CNN7-6 SM7,8 R1 Arm Up/Down Motor NB R1ZINB SMC CNN2-67 CNN3567 CNN33-12 SM1 SMD CNN1-12 SMD CNN7-1 – CNN7-6 SM7,8 R1 Arm Up/Down Motor TDA R1ZTDA SMC CNN2-68 CNN3568 CNN33-13 SM1 SMD CNN1-13 SMD CNN7-1 – CNN7-6 SM7,8 R1 Arm Up/Down Motor TDB R1ZTDB SMC CNN2-69 CNN3569 CNN33-14 SM1 SMD CNN1-14 SMD CNN7-1 – CNN7-6 SM7,8 R1 Arm Up/Down Motor Hold R1ZHOLD SMC CNN2-70 CNN3570 CNN33-15 SM1 SMD CNN1-15 SMD CNN7-1 – CNN7-6 SM7,8 R1 Arm U/D R1AEHOME SMC CNN 917 CNN 317 CNN 17-12 CNN17-10 +5V CNN17-11 GND R1 Arm U/D R1AEMIR SMC CNN 918 CNN 318 CNN 17-9 CNN17-7 +5V CNN17-8 GND R1 Arm U/D R1ACLASH SMC CNN 919 CNN 319 CNN 17-16 CNN17-14 +5V CNN17-15 GND R1 Arm Rotation Motor and Sensors R1 Arm Rotate Motor NA R1HINA SMC CNN2-61 CNN3561 CNN33-16 SM1 SMD CNN1-16 SMD CNN7-7 – CNN7-12 SM7,8 R1 Arm Rotate Motor NB R1HINB SMC CNN2-62 CNN3562 CNN33-17 SM1 SMD CNN1-17 SMD CNN7-7 – CNN7-12 SM7,8 R1 Arm Rotate Motor TDA R1HTDA SMC CNN2-63 CNN3563 CNN33-18 SM1 SMD CNN1-18 SMD CNN7-7 – CNN7-12 SM7,8 R1 Arm Rotate Motor TDB R1HTDB SMC CNN2-64 CNN3564 CNN33-19 SM1 SMD CNN1-19 SMD CNN7-7 – CNN7-12 SM7,8 R1 Arm Rotate Motor Hold R1HHOLD SMC CNN2-65 CNN3565 CNN33-20 SM1 SMD CNN1-20 SMD CNN7-7 – CNN7-12 SM7,8 R1 Arm Rotation R1AAHOME SMC CNN 915 CNN 315 CNN 17-6 CNN17-4 +5V CNN17-5 GND R1 Arm Rotation R1AAMIR SMC CNN 916 CNN 316 CNN 17-3 CNN17-1 +5V CNN17-2 GND +5V VCCR1AAUTOZERO SMC CNN 959 CNN3-59 CNN12-13 R1A Probe LLS Auto Zero R1AAUTOZERO SMC CNN 960 CNN3-17 CNN12-14 R1A Probe LLS (+) Differential VCCR1ALVL SMC CNN 955 CNN3-55 CNN12-15 R1A Probe LLS (-) Differential R1ALVL SMC CNN 956 CNN3-56 CNN12-16 R1A Probe LLS (+) Integral VCCR1ALVLOK SMC CNN 957 CNN3-57 CNN12-17 R1A Probe LLS (-) Integral R1ALVLOK SMC CNN 958 CNN3-58 CNN12-18 R1 Pressure Sensor Sig POS R1ASENSPOS PM CNN 103 CNN 6-3 CNN 25-7 +12V CNN 25-8 +/12V CNN 25-9 – 12V CNN 25-10 NC CNN 25-11 R1 Pressure Sensor Sig neg R1ASENSPOS PM CNN 104 CNN 6-4 CNN 25-12 LLS BD CNN 2 TO R1 Probe Liquid Level Sense Schematic Diagrams Reagent 1 Arm Motors and Sensors Link 7A_9021a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Reagent 1 Syringe Pump Motors, Sensors and Valves Reagent 1 Syringe Pump Motor and Sensor Cable Trace Description Card Cage Board Code Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable Reagent 1 Syringe Pump Motor and Sensors R1 Pump Motor NA R1PINA SMC CNN2-41 CNN35-41 CNN33-56 SM1 SMD CNN1-56 SMD SM11,12 CNN11-7 – CNN11-12 R1 Pump Motor NB R1PINB SMC CNN2-42 CNN35-42 CNN33-57 SM1 SMD CNN1-57 SMD SM11,12 CNN11-7 – CNN11-12 R1 Pump Motor TDA R1PTDA SMC CNN2-43 CNN35-43 CNN33-58 SM1 SMD CNN1-58 SMD SM11,12 CNN11-7 – CNN11-12 R1 Pump Motor TDB R1PTDB SMC CNN2-44 CNN35-44 CNN33-59 SM1 SMD CNN1-59 SMD SM11,12 CNN11-7 – CNN11-12 R1 Pump Motor Hold R1PHOLD SMC CNN2-45 CNN35-45 CNN33-60 SM1 SMD CNN1-60 SMD SM11,12 CNN11-7 – CNN11-12 R1 Pipettor R1PHOME SMC CNN 9-9 CNN 3-9 CNN 11-15 CNN11-13 +5V CNN11-14 GND R1 Pipettor R1PMIR SMC CNN 910 CNN 3-10 CNN 11-12 CNN11-10 +5V CNN11-11 GND R1 Pipettor R1PDLMT SMC CNN 911 CNN 3-11 CNN 11-18 CNN11-16 +5V CNN11-17 GND Pumps, Valves and Flow Sensor Probe Wash Pump (Inner) PWPUP AC/DC AC/DC CNN1-7 CNN34-7 CNN32-7 AD1 AC/DC CNN1-7 AC/DC CNN3-1 AD2,3 Probe Wash Pump (Inner) PWPDOWN AC/DC AC/DC CNN1-8 CNN34-8 CNN32-8 AD1 AC/DC CNN1-8 AC/DC CNN3-2 AD2,3 Probe Wash Pump (Inner) PWPBK AC/DC AC/DC CNN1-9 CNN34-9 CNN32-9 AD1 AC/DC CNN1-9 AC/DC CNN3-3 AD2,3 Probe Wash Pump (Inner) PWPUP AC/DC AC/DC CNN7-7 CNN2-7 CNN13-11 +5V CNN13-12 GND CNN13-13 Probe Wash Pump (Inner) PWPDOWN AC/DC AC/DC CNN7-8 CNN2-8 CNN13-14 +5V CNN13-15 GND CNN13-16 Reagent IR1PV Pump 1A (R1) MAG Valve (Valve ON) AC/DC AC/DC CNN34-43 CNN1-43 CNN32-29 AD1 AC/DC CNN1-29 Reagent IR1PV FB Pump 1A (R1) (SENS17 FB) MAG Valve (Valve ON) FB SM AC/DC SM CNN4-52 AC/DC CNN8-52 CNN32-53 AD1 AC/DC CNN1-53 DIW Supply Pump 2 (Outer Wall) WPMP AC/DC AC/DC CNN34-29 CNN1-29 CNN32-25 AD1 DIW Supply Pump 2 (Outer Wall) FB WPMP FB (SENS27) SM AC/DC SM CNN4-64 AC/DC CNN8-64 CNN32-49 DIW Valve 2 (R Probe Outer Wall) ERPV AC/DC AC/DC CNN34-34 CNN1-34 CNN32-36 DIW Valve 2 (R Probe Outer Wall) FB ERPV FB (SENS 7) SM AC/DC SM CNN4-43 AC/DC CNN8-43 CNN32-60 Reagent Wash Cup Flow Sensor RPFS SM AC/DC SENS30 SENS31 SENS32 CNN 14 –27 28 29 CNN38-4 +5V CNN38-5 GND CNN38-6 CNN37 Schematic Diagrams Reagent 1 Syringe Pump Motors and Valves AC/DC CNN8-3 CNN8-4 AD8 AC/DC CNN1-25 AC/DC CNN7-3 CNN7-6 AD6,7 AD1 AC/DC CNN1-36 AC/DC CNN9-5 CNN9-6 AD9 AD1 AC/DC CNN1-60 Link 7A_9028a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Reagent 2 Arm Motor and Sensor C4000 Reagent 2 Arm Motor and Sensor Cable Trace Description Code Card Cage Board Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable Reagent 2 Arm Up/Down Motor and Sensors R2 Arm Up/Down Motor NA R2ZINA SMC CNN3-6 CNN36-6 CNN33-21 SM1 SMD CNN1-21 SMD CNN8-1 – CNN8-6 SM7,8 R2 Arm Up/Down Motor NB R2ZINB SMC CNN3-7 CNN36-7 CNN33-22 SM1 SMD CNN1-22 SMD CNN8-1 – CNN8-6 SM7,8 R2 Arm Up/Down Motor TDA R2ZTDA SMC CNN3-8 CNN36-8 CNN33-23 SM1 SMD CNN1-23 SMD CNN8-1 – CNN8-6 SM7,8 R2 Arm Up/Down Motor TDB R2ZTDB SMC CNN3-9 CNN36-9 CNN33-24 SM1 SMD CNN1-24 SMD CNN8-1 – CNN8-6 SM7,8 R2 Arm Up/Down Motor Hold R2ZHOLD SMC CNN3-10 CNN3610 CNN33-25 SM1 SMD CNN1-25 SMD CNN8-1 – CNN8-6 SM7,8 R2 Arm U/D R2AEHOME SMC CNN 922 CNN 322 CNN 19-12 CNN19-10 +5V CNN19-11 GND R2 Arm U/D R2AEMIR SMC CNN 923 CNN 323 CNN 19-9 CNN19-7 +5V CNN19-8 GND R2 Arm U/D R2ACLASH SMC CNN 924 CNN 324 CNN 19-16 CNN19-14 +5V CNN19-15 GND Reagent 2 Arm Rotation Motor and Sensors R2 Arm Rotate Motor NA R2HINA SMC CNN3-1 CNN36-1 CNN33-26 SM1 SMD CNN1-26 SMD CNN8-7 – CNN8-12 SM8,9 R2 Arm Rotate Motor NB R2HINB SMC CNN3-2 CNN36-2 CNN33-27 SM1 SMD CNN1-27 SMD CNN8-7 – CNN8-12 SM8,9 R2 Arm Rotate Motor TDA R2HTDA SMC CNN3-3 CNN36-3 CNN33-28 SM1 SMD CNN1-28 SMD CNN8-7 – CNN8-12 SM8,9 R2 Arm Rotate Motor TDB R2HTDB SMC CNN3-4 CNN36-4 CNN33-29 SM1 SMD CNN1-29 SMD CNN8-7 – CNN8-12 SM8,9 R2 Arm Rotate Motor Hold R2HHOLD SMC CNN3-5 CNN36-5 CNN33-30 SM1 SMD CNN1-30 SMD CNN8-7 – CNN8-12 SM8,9 R2 Arm Rot R2AAHOME SMC CNN 920 CNN 320 CNN 19-6 CNN19-4 +5V CNN19-5 GND R2 Arm Rot R2AAMIR SMC CNN 9- CNN 3- CNN 19-3 21 21 CNN19-1 +5V CNN19-2 GND Liquid Level Sense +5V VCCR2AAUTOZERO SMC CNN 965 CNN3-65 CNN12-19 R2A Probe LLS Auto Zero R2AAUTOZERO SMC CNN 966 CNN3-22 CNN12-20 R2A Probe LLS (+) Differential VCCR2ALVL SMC CNN 961 CNN3-61 CNN12-21 R2A Probe LLS (-) Differential R2ALVL SMC CNN 962 CNN3-62 CNN12-22 R2A Probe LLS (+) Integral VCCR2ALVLOK SMC CNN 963 CNN3-63 CNN12-23 R2A Probe LLS (-) Integral R2ALVLOK SMC CNN 964 CNN3-64 CNN12-24 R2 Pressure Sensor Sig POS R2ASENSPOS PM CNN 105 CNN 6-5 CNN 25-13 +12V CNN 25-14+/12V CNN 25-15 – 12V CNN 25-16 NC CNN 25-17 R2 Pressure Sensor Sig neg R2ASENSPOS PM CNN 106 CNN 6-6 CNN 25-18 Level Sense Board CNN 3 to R2 Probe Schematic Diagrams Reagent 2 Arm Motors and Sensors Link 7A_9022a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Reagent 2 Syringe Pump Motors, Sensors and Valves C4000 Reagent 2 Syringe Pump Motor and Sensor Cable Trace Description Code Card Cage Back Plane Cable Card Cage Board CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable Reagent 2 Syringe Pump Motor and Sensors R2 Pump Motor NA R2PINA SMC CNN2-51 CNN3551 CNN33-61 SM1 SMD CNN1-61 SMD SM11,12 CNN12-1 – CNN12-6 R2 Pump Motor NB R2PINB SMC CNN2-52 CNN3552 CNN33-62 SM1 SMD CNN1-62 SMD SM11,12 CNN12-1 – CNN12-6 R2 Pump Motor TDA R2PTDA SMC CNN2-53 CNN3553 CNN33-63 SM1 SMD CNN1-63 SMD SM11,12 CNN12-1 – CNN12-6 R2 Pump Motor TDB R2PTDB SMC CNN2-54 CNN3554 CNN33-64 SM1 SMD CNN1-64 SMD SM11,12 CNN12-1 – CNN12-6 R2 Pump Motor Hold R2PHOLD SMC CNN2-55 CNN3555 CNN33-65 SM1 SMD CNN1-65 SMD SM11,12 CNN12-1 – CNN12-6 R2 Pipettor R2PHOME SMC CNN 912 CNN 3-12 CNN 11-24 CNN11-22 +5V CNN11-23 GND R2 Pipettor R2PMIR SMC CNN 913 CNN 3-13 CNN 11-21 CNN11-19 +5V CNN11-20 GND R2 Pipettor R2PDLMT SMC CNN 914 CNN 3-14 CNN 11-27 CNN11-25 +5V CNN11-26 GND Pumps, Valves and Flow Sensors Probe Wash Pump (Inner) PWPUP AC/DC AC/DC CNN1-7 CNN34-7 CNN32-7 AD1 AC/DC CNN1-7 AC/DC CNN3-1 AD2,3 Probe Wash Pump (Inner) PWPDOWN AC/DC AC/DC CNN1-8 CNN34-8 CNN32-8 AD1 AC/DC CNN1-8 AC/DC CNN3-2 AD2,3 Probe Wash Pump (Inner) PWPBK AC/DC AC/DC CNN1-9 CNN34-9 CNN32-9 AD1 AC/DC CNN1-9 AC/DC CNN3-3 AD2,3 Probe Wash Pump (Inner) PWPUP AC/DC AC/DC CNN7-7 CNN2-7 CNN13-11 +5V CNN13-12 GND CNN13-13 Probe Wash Pump (Inner) PWPDOWN AC/DC AC/DC CNN7-8 CNN2-8 CNN13-14 +5V CNN13-15 GND CNN13-16 Reagent IR2PV Pump 2A (R2) MAG Valve (Valve ON) AC/DC AC/DC CNN1-44 CNN3444 CNN32-30 AD1 AC/DC CNN1-30 AC/DC CNN8-5 CNN8-6 AD8 Reagent IR2PV FB Pump 2A (R2) (SENS17 MAG Valve FB) SM AC/DC SM AC/DC CNN8-53 CNN4-53 CNN32-54 AD1 AC/DC CNN1-54 (Valve ON) FB DIW Supply Pump 2 (Outer Wall) WPMP AC/DC AC/DC CNN1-29 CNN3429 CNN32-25 DIW Supply Pump 2 (Outer Wall) FB WPMP FB (SENS27) SM AC/DC SM AC/DC CNN8-64 CNN4-64 CNN32-49 DIW Valve 2 (R Probe Outer Wall) ERPV AC/DC AC/DC CNN1-34 CNN3434 DIW Valve 2 (R Probe Outer Wall) FB ERPV FB (SENS 7) SM AC/DC SM AC/DC CNN8-43 Reagent Wash Cup Flow Sensor RPFS SM AC/DC SENS30 SENS31 SENS32 CNN1427 28 29 AD1 AC/DC CNN1-25 AC/DC CNN7-3 CNN7-6 AD6,7 CNN32-36 AD1 AC/DC CNN1-36 AC/DC CNN9-5 CNN9-6 AD9 CNN4-43 CNN32-60 AD1 AC/DC CNN1-60 CNN37 CNN38-4 +5V CNN38-5 GND CNN38-6 Schematic Diagram Reagent 2 Syringe Pump Motors, Sensors and Valves Link 7A_9023a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Reagent Carousel Motor and Sensors C4000 Reagent Carousel Motor and Sensor Cable Trace Description Code Card Cage Board Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable Reagent Carousel, Inner and Outer Motors and Sensors Reagent Inner Carousel Motor CW RICCW SMC CNN353 CNN3653 CNN33-3 SM1 SMD CNN1-3 SMD CNN4-1 – SM4,5 CNN4-5 Reagent Inner Carousel Motor CCW RICCCW SMC CNN354 CNN3654 CNN33-4 SM1 SMD CNN1-4 SMD CNN4-1 – SM4,5 CNN4-5 Reagent Outer Carousel Motor CW ROCCW SMC CNN355 CNN3655 CNN33-5 SM1 SMD CNN1-5 SMD SM4,5 CNN5-1 – CNN5-5 Reagent Outer Carousel Motor CCW ROCCCW SMC CNN356 CNN3656 CNN33-6 SM1 SMD CNN1-6 SMD SM4,5 CNN5-1 – CNN5-5 Rgt Inner Crsl R1CIHOME SMC CNN935 CNN 335 CNN 16-6 CNN16-4 +5V CNN16-5 GND Rgt Inner Crsl R1CI PITCH SMC CNN 936 CNN 336 CNN 16-3 CNN16-1 +5V CNN16-2 GND Rgt Inner Crsl R1CIPOS SMC CNN 937 CNN 337 Rgt Outer Crsl R1COHOME SMC CNN 938 CNN 338 CNN 16-12 CNN16-10 +5V CNN16-11 GND Rgt Outer Crsl R1COPITCH SMC CNN 939 CNN 339 CNN 16-9 CNN16-7 +5V CNN16-8 GND Rgt Outer Crsl R1COPOS SMC CNN 940 CNN 340 Covers, Switches and Barcode Readers Reagent Load Door Open/Close AUXAC1 AC/DC CNN113 CNN3413 CNN32-10 AD1 AC/DC CNN1-10 AC/DC CNN5-4 CNN5-6 AD4,5 Reagent Load Door Open/Close AUXAC1 AC/DC CNN114 CNN3414 CNN32-11 AD1 AC/DC CNN1-11 AC/DC CNN5-4 CNN5-6 AD4,5 Reagent Load Door Open/Close AUXAC1 AC/DC CNN115 CNN3415 CNN32-12 AD1 AC/DC CNN1-12 AC/DC CNN5-4 CNN5-6 AD4,5 Reagent Supply Temp Sensor R1CTEMP- AC/DC CNN 786 CNN2-86 CNN16-13 Reagent R1CTEMP+ AC/DC CNN 7- CNN2-85 CNN16-14 Supply Temp Sensor 85 Reagent RCVROK Supply Cover Present Sensor AC/DC Outer Rotate Switch Inner Rotate Switch CNN 759 CNN2-59 CNN16-17 CNN16-15 +5V CNN16-16 GND AC/DC CNN4- CNN16-20 CNN16-18 +5V CNN16-19 GND AC/DC CNN4- CNN16-23 CNN16-21 +5V CNN16-22 GND Bottle Sensor BTTLSNS AC/DC CNN 795 CNN2-95 CNN16-26 CNN16-24 +5V CNN16-25 GND Cup Sensor CUPSNSRO AC/DC CNN 796 CNN2-96 CNN16-29 CNN16-27 +5V CNN16-28 GND Cover Open/Close Sensor Open COVER OPEN CNN2-10 CNN16-37 CNN16-35 +5V CNN16-36 GND Cover Open/Close Sensor Close COVER CLOSE CNN2-9 CNN16-40 CNN16-38 +5V CNN16-39 GND Reagent Barcode Reader TXD2 DAQ CNN 1211 CNN7-11 CNN20-1 Reagent Barcode Reader RTS2 DAQ CNN 1213 CNN7-13 CNN20-2 Reagent Barcode Reader RXD2 DAQ CNN 1212 CNN7-12 CNN20-4 Reagent Barcode Reader CTS2 DAQ CNN 1214 CNN7-14 CNN20-5 Reagent Barcode Reader TRIGGER2 DAQ CNN 1215 CNN7-15 CNN20-8 Rotate Switch ON ISLUILED1(RCLED) AC/DC CNN 736 CNN2-36 CNN30-2 CNN30-1 +5V Open/Close Switch ON ISLUISLED2(RCLED) AC/DC CNN 738 CNN2-38 CNN30-4 CNN30-3 +5V Rotate Switch ISSW(RCSW) AC/DC CNN 791 CNN2-91 CNN30-6 CNN30-5 GND Open/Close Switch ISLSW(RCSW) AC/DC CNN 792 CNN2-92 CNN30-8 CNN30-7 GND Schematic Diagram Reagent Carousel Motors and Sensors Link 7A_9024a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Sample Arm Motors and Sensors C4000 Sample Arm Motor and Sensor Cable Trace Description Card Cage Board Code Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Driver Box Mixer Bd Cable Cable Pins Sample Arm Up/Down Motor and Sensors Sample Arm Up/Down Motor CW SAZCW SMC CNN3-63 CNN36-63 CNN33-7 SM1 SMD CNN1-7 SMD CNN3-1 – CNN3-5 SM2,3 Sample Arm Up/Down Motor CCW SAZCCW SMC CNN3-64 CNN36-64 CNN33-8 SM1 SMD CNN1-8 SMD CNN3-1 – CNN3-5 SM2,3 Sample U/D SAEHOME SMC CNN 9-46 CNN 3-46 CNN 1512 Sample U/D SAEMIR SMC CNN 9-47 CNN 3-47 CNN 15-9 Sample U/D SACLASH SMC CNN 9-48 CNN 3-48 CNN 1516 Sample Arm Rotation Motor and Sensors Sample Arm Rotate Motor CW SAHCW SMC CNN3-61 CNN36-61 CNN33-9 SM1 SMD CNN1-9 SMD CNN6-1 – CNN6-5 SM6 Sample Arm Rotate Motor CCW SAHCCW SMC CNN3-62 CNN36-62 CNN33-10 SM1 SMD CNN1-10 SMD CNN6-1 – CNN6-5 SM6 Sample Rotation SAAHOME SMC CNN 9-44 CNN 3-44 CNN 15-6 Sample Rotation SAAMIR SMC CNN 9-45 CNN 3-45 CNN 15-3 +5V VCCSAAAUTOZERO SMC CNN 9-53 CNN3-53 CNN12-7 S Probe LLS Austo Zero SAAUTOZERO SMC CNN 9-54 CNN3-46 CNN12-8 S Probe LLS (+) Differential VCCSALVL SMC CNN 9-49 CNN3-49 CNN12-9 S Probe LLS (-) Differential SALVL SMC CNN 9-50 CNN3-50 CNN12-10 S Probe LLS (+) Integral VCCSALVLOK SMC CNN 9-51 CNN3-51 CNN12-11 S Probe LLS (-) Integral SALVLOK SMC CNN 9-52 CNN3-52 CNN12-12 SALVLGU+5V SMC CNN 9-68 CNN3-68 CNN12-25 SALVLGU SMC CNN 9-67 CNN3-67 CNN12-26 S Pressure Sensor Sig POS SASEN1 POS PM CNN 10-1 CNN6-1 CNN 11-1 CNN25-5 Pressure Sensor Sig SASEN1 NEG PM CNN 10-2 CNN6-2 CNN 11-2 CNN25-6 Liquid Level Sense Neg Sample TLLS SLLS PM CNN 1033 CNN 6-33 CNN25-19 LLS BD 5- Sample TLLS SLLSG PM CNN 1034 CNN 6-34 CNN25-20 LLS BD 5- CNN 12 TO LLS Bd to LLS Bd TLLS-SR Schematic Diagram Sample Arm Motors and Sensors Link 7A_9025a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Sample Syringe Pump Motor and Sensors C4000 Sample Syringe Pump Motor and Sensor Cable Trace Description Code Card Cage Card Cage Back CNN Board Board Plane Cable Cable CNN Board Cable Driver Box Cable SMD or SMD or AC/DC or AC/DC or Mixer Bd Mixer Bd I/F Signal Cable Pins Cable Pins Driver Box Cable Sample Syringe Pump Motor and Sensors Sample Pump Motor NA SPINA SMC CNN2-31 CNN35-31 CNN33-51 SM1 SMD CNN1-51 SMD SM10,11 CNN11-1 – CNN11-6 Sample Pump Motor NB SPINB SMC CNN2-32 CNN35-32 CNN33-52 SM1 SMD CNN1-52 SMD SM10,11 CNN11-1 – CNN11-6 Sample Pump Motor TDA SPTDA SMC CNN2-33 CNN35-33 CNN33-53 SM1 SMD CNN1-53 SMD SM10,11 CNN11-1 – CNN11-6 Sample Pump Motor TDB SPTDB SMC CNN2-34 CNN35-34 CNN33-54 SM1 SMD CNN1-54 SMD SM10,11 CNN11-1 – CNN11-6 Sample Pump Motor Hold SPHOLD SMC CNN2-35 CNN35-35 CNN33-55 SM1 SMD CNN1-55 SMD SM10,11 CNN11-1 – CNN11-6 Sample Pipettor SPHOME SMC CNN 9-6 CNN 3-6 CNN 11-6 Sample Pipettor SPMIR SMC CNN 9-7 CNN 3-7 CNN 11-3 Sample Pipettor SPDLMT SMC CNN 9-8 CNN 3-8 CNN 11-9 Pumps, Valves and Sensors Probe Wash Pump (Inner) PWPUP AC/DC AC/DC CNN1-7 CNN34-7 CNN32-7 AD1 AC/DC CNN1-7 AC/DC CNN3-1 AD2,3 Probe Wash Pump (Inner) PWPDOWN AC/DC AC/DC CNN1-8 CNN34-8 CNN32-8 AD1 AC/DC CNN1-8 AC/DC CNN3-2 AD2,3 Probe Wash Pump (Inner) PWPBK AC/DC AC/DC CNN1-9 CNN34-9 CNN32-9 AD1 AC/DC CNN1-9 AC/DC CNN3-3 AD2,3 Probe Wash Pump (Inner) PWPUP AC/DC AC/DC CNN7-7 CNN2-7 CNN13-11 +5V CNN13-12 GND CNN13-13 Probe Wash Pump (Inner) PWPDOWN AC/DC AC/DC CNN7-8 CNN2-8 CNN13-14 +5V CNN13-15 GND CNN13-16 Sampling ISPV Pump A MAG Valve (Valve 1 & 2 ON) AC/DC AC/DC CNN1-42 CNN34-42 CNN32-28 AD1 AC/DC CNN1-28 AC/DC CNN8-1 CNN8-2 AD8 Sampling ISPV FB Pump A MAG (SENS16 Valve (Valve 1 FB) SM AC/DC SM AC/DC CNN4-51 CNN8-51 CNN32-52 AD1 AC/DC CNN1-52 & 2 ON) FB DIW Supply Pump 2 (Outer Wall) WPMP AC/DC CNN34-29 CNN32-25 DIW Supply Pump 2 (Outer Wall) FB WPMP FB (SENS27) SM AC/DC SM AC/DC CNN4-64 CNN8-64 CNN32-49 DIW Valve 1 (S Probe Outer Wall) ESPV AC/DC CNN34-33 DIW Valve 1 (S Probe Outer Wall) FB ESPV FB (SENS 6) Sample Wash Cup Flow Sensor SPFS AC/DC CNN1-29 AD1 AC/DC CNN1-25 AC/DC CNN7-3 CNN7-6 AD6,7 CNN32-35 AD1 AC/DC CNN1-35 AC/DC CNN9-3 CNN9-4 AD9 SM AC/DC SM AC/DC CNN4-42 CNN8-42 CNN32-59 AD1 AC/DC CNN1-59 SM AC/DC SENS 30 SENS 31 SENS 32 CNN38-1 +5V CNN38-2 GND CNN38-3 AC/DC CNN1-33 SM AC/DC CNN37 CNN 1427 CNN 1428 CNN 1429 Schematic Diagram Sample Syringe Pump Motors and Sensors Link 7A_9026a.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Stirrer Motors and Sensors Stirrer Motor and Sensor Cable Trace Description Code Card Cage Board Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD or AC/DC or Mixer Bd I/F Signal Cable Pins SMD or AC/DC or Mixer Bd Cable Pins Driver Box Cable Stirrer Elevator Stirrer Arm CW Signal (UP) STRZ UP AC/DC CNN1-4 CNN34-4 CNN32-4 AD1 AC/DC CNN1-4 AC/DC CNN2-4 – CNN2-6 AD2,3 (to STRE) Stirrer Arm CCW Signal (DOWN) STRZ DW AC/DC CNN1-5 CNN34-5 CNN32-5 AD1 AC/DC CNN1-5 AC/DC CNN2-4 – CNN2-6 AD2,3 (to STRE) Stirrer Arm Brake Signal STRZ BK AC/DC CNN1-6 CNN34-6 CNN32-6 AD1 AC/DC CNN1-6 AC/DC CNN2-4 – CNN2-6 AD2,3 (to STRE) Stirrer (Sensor) STEUP AC/DC CNN7-4 CNN 2-4 CNN 18-3 CNN18-1 +5V CNN18-2 GND (to SE-S) Stirrer (Sensor) STEDOWN AC/DC CNN 7-5 CNN 2-5 CNN 18-6 CNN18-4 +5V CNN18-5 GND (to SE-S) Stirrer (Sensor) STEDOK AC/DC CNN 7-6 CNN 2-6 CNN 18-9 CNN18-7 +5V CNN18-8 GND (to SE-S) Stirrer 1 Rotate Motor NA STR1HINA SMC CNN2-11 CNN35-11 CNN33-31 SM1 SMD CNN1-31 SMD CNN9-1 – CNN9-6 SM8,9 (to STRA) Stirrer 1 Rotate Motor NB STR1HINB SMC CNN2-12 CNN35-12 CNN33-32 SM1 SMD CNN1-32 SMD CNN9-1 – CNN9-6 SM8,9 (to STRA) Stirrer 1 Rotate Motor TDA STR1HTDA SMC CNN2-13 CNN35-13 CNN33-33 SM1 SMD CNN1-33 SMD CNN9-1 – CNN9-6 SM8,9 (to STRA) Stirrer 1 Rotate Motor TDB STR1HTDB SMC CNN2-14 CNN35-14 CNN33-34 SM1 SMD CNN1-34 SMD CNN9-1 – CNN9-6 SM8,9 (to STRA) Stirrer 1 Rotate Motor Hold STR1HHOLD SMC CNN2-15 CNN35-15 CNN33-35 SM1 SMD CNN1-35 SMD CNN9-1 – CNN9-6 SM8,9 (to STRA) Stirrer 1 Rot Hm STA1HOME SMC CNN 9-4 CNN 3-4 CNN 18-12 CNN18-10 +5V CNN1811GND (to SE-S) to SE-S Stirrer 1 SENS1 SM AC/DC CNN 8- CNN4-37 CNN18-15 (to To STRC Stirrer 1 Rotate Cuvette Sense 37 CNN18-13 +5V CNN18-14 GND Piezo Stirrer STRC) CNN26-1 +5V CNN26-2 +5V Stirrer 1 ON ST1ON AC/DC CNN 723 CNN2-23 CNN26-3 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 1 Scan ST1SCAN AC/DC CNN 724 CNN2-24 CNN26-4 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 1 Clear ST1CLS AC/DC CNN 725 CNN2-25 CNN26-5 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 1 Busy 0 Sig ST1BUSY0 AC/DC CNN 726 CNN2-26 CNN26-6 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 1 Busy 1 Sig ST1BUSY1 AC/DC CNN 727 CNN2-27 CNN26-7 STR1 Mixer CN1 Mixer CN2 STR2 to STRD Stirrer 2 Rotate Motor NA STR2HINA SMC CNN2-16 CNN35-16 CNN33-36 SM1 SMD CNN1-36 SMD CNN9-7 – CNN9-12 SM8 (to STRA) Stirrer 2 Rotate Motor NB STR2HINB SMC CNN2-17 CNN35-17 CNN33-37 SM1 SMD CNN1-37 SMD CNN9-7 – CNN9-12 SM8 (to STRA) Stirrer 2 Rotate Motor TDA STR2HTDA SMC CNN2-18 CNN35-18 CNN33-38 SM1 SMD CNN1-38 SMD CNN9-7 – CNN9-12 SM8 (to STRA) Stirrer 2 Rotate Motor TDB STR2HTDB SMC CNN2-19 CNN35-19 CNN33-39 SM1 SMD CNN1-39 SMD CNN9-7 – CNN9-12 SM8 (to STRA) Stirrer 2 Rotate Motor Hold STR2HHOLD SMC CNN2-20 CNN35-20 CNN33-40 SM1 SMD CNN1-40 SMD CNN9-7 – CNN9-12 SM8 (to STRA) Stirrer 2 Rot Hm STA2HOME SMC CNN 9-5 CNN 3-5 CNN 18-18 CNN18-16 +5V CNN18-17 GND (to SE-S) Stirrer 2 Rotate Cuvette Sense SENS2 SM AC/DC CNN 838 CNN4-38 CNN18-21 CNN18-20 +5V CNN18-21 GND (to STRC) Stirrer 2 ON ST2ON AC/DC CNN 728 CNN2-28 CNN26-8 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 2 Scan ST2SCAN AC/DC CNN 729 CNN2-29 CNN26-9 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 2 Clear ST2CLS AC/DC CNN 730 CNN2-30 CNN26-10 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 2 Busy 0 Sig ST2BUSY0 AC/DC CNN 731 CNN2-31 CNN26-11 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) Stirrer 2 Busy 1 Sig ST2BUSY1 AC/DC CNN 732 CNN2-32 CNN26-12 STR1 Mixer CN1 Mixer CN2 STR2 (to STRD) AC/DC CNN1-29 CNN34-29 CNN32-25 AD1 AC/DC CNN1-25 AC/DC CNN7-3 AD6,7 Stirrer 2 Stirrer Wash Valve and Flow Sensor Signal DIW Supply Pump 2 WPMP AC/DC (Outer Wall) CNN7-6 DIW Supply Pump 2 (Outer Wall) FB WPMP FB (SENS27) SM AC/DC SM AC/DC CNN8-64 CNN4-64 CNN32-49 DIW Valve 3 Stirrer Wash Valve SWV AC/DC AC/DC CNN1-35 CNN34-35 CNN32-37 AD1 AC/DC CNN1-37 Stirrer Wash Valve Signal FB SWV FB (SENS8) SM AC/DC SM AC/DC CNN8-44 CNN4-44 CNN32-61 AD1 AC/DC CNN1-61 Stirrer Line Flow Signal STFS SM AC/DC SENS 30 SENS 31 SENS 32 SM AC/DC CNN37- CNN38-9 CNN38-7 +5V CNN38-8 GND Power Supply CNN1427 CNN 1428 CNN 1429 STRD Driver Box STR3,4 Schematic Diagram Stirrer Motor and Sensors AC/DC CNN9-7 – CNN9-8 AD9 Mixer Board CN3 CN4 Link 7A_9027b.pdf ARCHITECT c4000 Service and Support Manual (Version 204733-107) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Individual Motor and Sensor Signal Trace Links AC/DC Driver Signal Trace AC/DC Sensors Sensors, Switches, Valves Stepper Motor Controller Sensors Stepper Motor Signal Trace ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. AC/DC Driver Signal Trace Description Code Card Cage Board Card Cage Back Plane Cable CNN Bd Cable CNN Bd Cable Driver Box Cable AC/DC Driver Board I/F Signal Drive Signal Output Cell Wash CW Signal WS UP AC/DC CNN1-1 CNN34-1 CNN32-1 AD1 CNN1-1 Cell Wash CCW Signal WS DWN AC/DC CNN1-2 CNN34-2 CNN32-2 AD1 CNN1-2 Cell Wash Brake Signal WS BK AC/DC CNN1-3 CNN34-3 CNN32-3 AD1 CNN1-3 Stirrer Arm CW Signal STRZ UP AC/DC CNN1-4 CNN34-4 CNN32-4 AD1 CNN1-4 Stirrer Arm CCW Signal STRZ DW AC/DC CNN1-5 CNN34-5 CNN32-5 AD1 CNN1-5 Stirrer Arm Brake STRZ BK Signal AC/DC CNN1-6 CNN34-6 CNN32-6 AD1 CNN1-6 Probe Inner Wall Wash Pump CW Signal PWP UP AC/DC CNN1-7 CNN34-7 CNN32-7 AD1 CNN1-7 Probe Inner Wall Wash Pump CCW Signal PWP DWN AC/DC CNN1-8 CNN34-8 CNN32-8 AD1 CNN1-8 Probe Inner Wall Wash Pump Brake Signal PWP BK AC/DC CNN1-9 CNN34-9 CNN32-9 AD1 CNN1-9 Reagent Load AUXAC1 Door Open/Close AC/DC CNN1-13 CNN34-13 CNN32-10 AD1 CNN1-10 Reagent Load AUXAC1 Door Open/Close AC/DC CNN1-14 CNN34-14 CNN32-11 AD1 CNN1-11 Reagent Load AUXAC1 Door Open/Close AC/DC CNN1-15 CNN34-15 CNN32-12 AD1 CNN1-12 CNN32-13 AD1 CNN1-13 NC CNN2-1 – CNN2-3 CNN3-1 – CNN3-3 CNN5-4 CNN56 AD4,5 CNN3-4 – CNN3-6 AD2,3 CNN4-1 – CNN4-3 AD4,5 WCP ON AC/DC CNN1-18 CNN34-18 CNN32-14 AD1 CNN1-14 Cell Wash Pump Brake Signal WCP BK AC/DC CNN1-19 CNN34-19 CNN32-15 AD1 CNN1-15 ISE Suction ICTP ON Pump CW Signal AC/DC CNN1-20 CNN34-20 CNN32-16 AD1 CNN1-16 ISE Suction Pump Brake Signal ICTP BK AC/DC CNN1-21 CNN34-21 CNN32-17 AD1 CNN1-17 ISE Ref Pump CW Signal ICTR ON AC/DC CNN1-22 CNN34-22 CNN32-18 AD1 CNN1-18 ISE Ref Pump CCW Signal ICTR BK AC/DC CNN1-23 CNN34-23 CNN32-19 AD1 CNN1-19 Detergent Pump CW Signal DTGP ON AC/DC CNN1-16 CNN34-16 CNN32-20 AD1 CNN1-20 Detergent Pump Brake Signal DTG BK AC/DC CNN1-17 CNN34-17 CNN32-21 AD1 CNN1-21 CNN32-22 AD1 CNN1-22 Cell Vacuum Pump ON Signal VP2 AC/DC CNN1-26 CNN34-26 CNN32-23 AD1 CNN1-23 CNN6-1 CNN62 Cell Dryer Pump ON Signal VP1 AC/DC CNN1-27 CNN34-27 CNN32-24 AD1 CNN1-24 CNN7-1 CNN74 Outer Wall Pump WPMP ON Signal AC/DC CNN1-29 CNN34-29 CNN32-25 AD1 CNN1-25 CNN7-3 CNN76 Auto Power OFF-ON Signal MC1 AC/DC CNN1-30 CNN34-30 CNN32-26 AD1 CNN1-26 CNN14-1 – CNN14-3 CNN32-27 AD1 CNN1-27 Sample Pump Valve 1 & 2 ON Signal ISPV AC/DC CNN1-42 CNN34-42 CNN32-28 AD1 CNN1-28 CNN8-1 CNN82 R1 Pump Valve IR1PV AC/DC CNN1-43 CNN34-43 CNN32-29 AD1 CNN1-29 CNN8-3 CNN8- NC AD2,3 CNN2-4 – CNN2-6 Cell Wash Pump CW Signal NC Driver Box Cable CNN4-4 – CNN4-6 CNN5-1 – CNN5-3 AD6,7 AD8 ON Signal 4 R2 Pump Valve ON Signal IR2PV AC/DC CNN1-44 CNN34-44 CNN32-30 AD1 CNN1-30 CNN8-5 CNN86 ISE Suction Pinch Valve ON Signal IAVP AC/DC CNN1-32 CNN34-32 CNN32-31 AD1 CNN1-31 CNN8-7 CNN88 Acid & Alkaline Detergent Valve 1 ON Signal DTG1V AC/DC CNN1-39 CNN34-39 CNN32-32 AD1 CNN1-32 CNN8-9 CNN810 Acid & Alkaline Detergent Valve 2 ON Signal DTG2V AC/DC CNN1-40 CNN34-40 CNN32-33 AD1 CNN1-33 CNN8-11 CNN812 Cell Wash Valve CWV AC/DC CNN1-38 CNN34-38 CNN32-34 AD1 CNN1-34 CNN9-1 CNN92 Sample Probe Valve ON Signal ESPV AC/DC CNN1-33 CNN34-33 CNN32-35 AD1 CNN1-35 CNN9-3 CNN94 R-Probe Outer Wash Valve ON Signal ERPV AC/DC CNN1-34 CNN34-34 CNN32-36 AD1 CNN1-36 CNN9-5 CNN96 Stirrer Wash Valve ON Signal SWV AC/DC CNN1-35 CNN34-35 CNN32-37 AD1 CNN1-37 CNN9-7 CNN98 DIW Supply Valve ON Signal WBSV AC/DC CNN1-36 CNN34-36 CNN32-38 AD1 CNN1-38 CNN9-9 CNN910 DIW Supply Pump (Peri Pump) ON PWV AC/DC CNN1-37 CNN34-37 CNN32-39 AD1 CNN1-39 CNN9-11 CNN912 ICT-Level Sensor ON Signal IREF AC/DC CNN1-49 CNN34-49 CNN32-40 AD1 CNN1-40 CNN10-1 CNN10-2 Buzzer ON Signal BUZZER AC/DC CNN1-48 CNN34-48 CNN32-41 AD1 CNN1-41 CNN10-3 CNN10-4 Spare 1 ON Signal (CRA) AUXDC1 AC/DC CNN1-50 CNN34-67 CNN32-43 AD1 CNN1-42 CNN10-5 CNN10-6 Spare 2 ON Signal AUXDC2 CNN34-68 CNN32-44 AD1 CNN1-43 CNN10-7 CNN10-8 CRA Door Open/Close 1 Signal AUTO1 AD1 CNN1-44 CNN10-9 CNN10-10 CRA Door Open/Close 2 Signal AUTO2 AD1 CNN1-45 NC AD1 CNN1-46 NC AD1 CNN1-47 NC AD1 CNN1-48 NC AD1 CNN1-49 NC AD1 CNN1-50 NC AD1 CNN1-51 Sample pump Valve 1 & 3 FB Signal ISPV FB CNN4-51 CNN32-52 AD1 CNN1-52 R1 Pump Valve FB Signal IR1PV FB CNN4-52 CNN32-53 AD1 CNN1-53 R2 Pump Valve FB Signal IR2PV FB CNN4-53 CNN32-54 AD1 CNN1-54 ISE Suction Pinch Valve FB Signal IAV FB CNN4-41 CNN32-55 AD1 CNN1-55 Acid & Alkaline Valve 1 FB Signal DTG1V FB CNN4-48 CNN32-56 AD1 CNN1-56 Acid & Alkaline Valve 2 FB Signal DTG2V FB CNN4-49 CNN32-57 AD1 CNN1-57 Cell Wash Valve FB Signal CWV FB CNN4-47 CNN32-58 AD1 CNN1-58 Sample Probe Outer Wash Valve FB Signal ESPV FB CNN4-42 CNN32-59 AD1 CNN1-59 AD9 AD10 R-Probe Outer ERPV FB Wall Wash Valve FB Signal CNN4-43 CNN32-60 AD1 CNN1-60 Stirrer Wash Valve FB Signal SWV FB CNN4-44 CNN32-61 AD1 CNN1-61 DIW Supply Valve FB Signal WBSV FB CNN4-45 CNN32-62 AD1 CNN1-62 DIW Supply Pump (Peri) FB Signal PWV FB CNN4-46 CNN32-63 AD1 CNN1-63 ICT-Level Sensor FB Signal IREF FB CNN4-58 CNN32-64 AD1 CNN1-64 Buzzer FB Signal BUZZER FB CNN4-57 CNN32-65 AD1 CNN1-65 Spare 1 FB Signal AUXDC1 FB CNN4-55 CNN32-67 AD1 CNN1-66 Spare 2 FB Signal AUXDC2 FB CNN4-56 CNN32-68 AD1 CNN1-67 NC AD1 CNN1-68 GND (+12V) Feedback GND GND (+12V) AD1 CNN1-69 GND (+12V) Feedback GND GND (+12V) AD1 CNN1-70 GND GND AD1 CNN1-71 GND GND AD1 CNN1-72 GND GND AD1 CNN1-73 GND GND AD1 CNN1-74 GND GND AD1 CNN1-75 +5V Power Supply +5V AD1 CNN1-76 +5V Power Supply +5V AD1 CNN1-77 +5V Power Supply +5V AD1 CNN1-78 +5V Power Supply +5V AD1 CNN1-79 AD1 CNN1-80 +5V Power Supply ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. AC/DC Sensors Description Wash Station Stirrer Probe Wash Pump AC1 Detergent Pump Cuvette Wash Pump ISE Aspir Pump ISE Ref Pump Code Card Cage Board Card Cage Back Plane Cable CNN Board CNN Board WSUP AC/DC CNN 7-1 CNN 2-1 CNN 18-22 +5V CNN 18-23 GND CNN 18-24 WSDOWN AC/DC CNN 7-2 CNN 2-2 CNN 18-25 +5V CNN 18-26 GND CNN 18-27 WSDOWNOK AC/DC CNN 7-3 CNN 2-3 CNN 18-28 +5V CNN 18-29 GND CNN 18-30 STEUP AC/DC CNN 7-4 CNN 2-4 CNN 18-1 +5V CNN 18-2 GND CNN 18-3 STEDOWN AC/DC CNN 7-5 CNN 2-5 CNN 18-4 +5V CNN 18-5 GND CNN 18-6 STEDOK AC/DC CNN 7-6 CNN 2-6 CNN 18-7 +5V CNN 18-8 GND CNN 18-9 PWPUP AC/DC CNN 7-7 CNN 2-7 CNN 13-11 +5V CNN 13-12 GND CNN 13-13 PWPDOWN AC/DC CNN 7-8 CNN 2-8 CNN 13-14 +5V CNN 13-15 GND CNN 13-16 AC1UP AC/DC CNN 7-9 AC1DOWN AC/DC CNN 7-10 AC1OK AC/DC CNN 7-11 DPUP AC/DC CNN 7-12 CNN 2-12 CNN 14-1 +5V CNN 14-2 GND CNN 14-3 DPDOWN AC/DC CNN 7-13 CNN 2-13 CNN 14-4 +5V CNN 14-5 GND CNN 14-6 CWPUP AC/DC CNN 7-14 CNN 2-14 CNN 14-16 +5V CNN 14-17 GND CNN 14-18 CWPDOWN AC/DC CNN 7-15 CNN 2-15 CNN 14-19 +5V CNN 14-20 GND CNN 14-21 ISPUP AC/DC CNN 7-16 CNN 2-16 CNN 25-21 +5V CNN 25-22 GND CNN 25-23 ISPDOWN AC/DC CNN 7-17 CNN 2-17 CNN 25-24 +5V CNN 25-25 GND CNN 25-26 IMPUP AC/DC CNN 7-18 CNN 2-18 CNN 25-27 +5V CNN 25-28 GND CNN 25-29 ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Sensors, Switches, Valves Description Code Card Cage Board Card Cage Backplane CNN Board CNN Board +5V VCCSAAAUTOZERO SMC 9-53 CNN3-53 CNN12-7 S Probe LLS Austo Zero SAAUTOZERO SMC 9-54 CNN3-46 CNN12-8 S Probe LLS (+) Differential VCCSALVL SMC 9-49 CNN3-49 CNN12-9 S Probe LLS (-) Differential SALVL SMC 9-50 CNN3-50 CNN12-10 S Probe LLS (+) Integral VCCSALVLOK SMC 9-51 CNN3-51 CNN12-11 S Probe LLS (-) Integral SALVLOK SMC 9-52 CNN3-52 CNN12-12 +5V VCCR1AAUTOZERO SMC 9-59 CNN3-59 CNN12-13 R1A Probe LLS Auto Zero R1AAUTOZERO SMC 9-60 CNN3-17 CNN12-14 R1A Probe LLS (+) Differential VCCR1ALVL SMC 9-55 CNN3-55 CNN12-15 R1A Probe LLS (-) Differential R1ALVL SMC 9-56 CNN3-56 CNN12-16 R1A Probe LLS (+) Integral VCCR1ALVLOK SMC 9-57 CNN3-57 CNN12-17 R1A Probe LLS (-) Integral R1ALVLOK SMC 9-58 CNN3-58 CNN12-18 +5V VCCR2AAUTOZERO SMC 9-65 CNN3-65 CNN12-19 R2A Probe LLS Auto Zero R2AAUTOZERO SMC 9-66 CNN3-22 CNN12-20 R2A Probe LLS (+) Differential VCCR2ALVL SMC 9-61 CNN3-61 CNN12-21 R2A Probe LLS (-) Differential R2ALVL SMC 9-62 CNN3-62 CNN12-22 R2A Probe LLS (+) Integral VCCR2ALVLOK SMC 9-63 CNN3-63 CNN12-23 R2A Probe LLS (-) Integral R2ALVLOK SMC 9-64 CNN3-64 CNN12-24 SALVLGU+5V SMC 9-68 CNN3-68 CNN12-25 SALVLGU SMC 9-67 CNN3-67 CNN12-26 CNN12-1,2 +15V CNN12-3,4 AGND CNN12-5,6 –15V +5V AC/DC +5V CNN13-1 AC/DC CNN13-2 Water Bath Heater ON/OFF IHTRON AC/DC 7-77 CNN2-77 CNN13-3 Water Bath Cycle Pump ON/OFF IPMON AC/DC 7-76 CNN2-76 CNN13-4 IFANON AC/DC 7-80 CNN2-80 CNN13-6 NC Water Bath Fan Cntrl ON/OFF AC/DC CNN13-5 Water Bath Mtr Vlv Open IMVOPEN AC/DC 7-78 CNN2-78 CNN13-7 Water Bath Mtr Vlv Clsd IMVSHUT AC/DC 7-79 CNN2-79 CNN13-8 Water Bath Fan Ctrl Thrm IVRSTR+ AC/DC 7-89 CNN2-89 CNN13-9 Water Bath Fan Ctrl Thrm IVRSTR- AC/DC 7-90 CNN2-90 +5V AC/DC GND CNN13-10 CNN13-11 AC/DC CNN13-12 Degasser Fail Detection DALARMV AC/DC 7-69 CNN2-69 CNN13-17 Degasser Fail Detection DALARM AC/DC 7-68 CNN2-68 CNN13-18 ISE Ref Solution ESMCC AC/DC 7-70 CNN2-70 CNN14-7 +5V CNN14-8 GND CNN14-9 Acid Detergent ESB3 AC/DC 7-72 CNN2-72 CNN14-10 +5V CNN14-11 GND CNN14-12 Alkaline Detergent ESB14 AC/DC 7-71 CNN2-71 CNN14-13 +5V CNN14-14 GND CNN14-15 Open Tank Upper Limit FFS2UPPERLIM AC/DC 7-63 CNN2-63 CNN14-22 GND AC/DC CNN14-23 Open Tank Low Level FFS4OPENL AC/DC 7-65 CNN2-65 CNN14-24 GND CNN14-25 Open Tank Full Level FFS3OPENH AC/DC 7-64 CNN2-64 CNN14-26 GND CNN14-27 LCW Full Detection FFS1LW AC/DC 7-62 CNN2-62 CNN14-28 GND CNN14-29 HCW Full Detection DTFHW AC/DC 7-67 CNN2-67 CNN14-30 GND CNN14-31 Spare Float SW FFS5AUXFSW AC/DC 7-66 CNN2-66 CNN14-32 GND CNN14-33 NC AC/DC CNN14-34 AC/DC Reagent Supply Temp Sensor R1CTEMP- AC/DC 7-86 CNN2-86 CNN16-13 Reagent Supply Temp Sensor R1CTEMP+ AC/DC 7-85 CNN2-85 CNN16-14 Reagent Supply Cover Present Sensor RCVROK AC/DC 7-59 CNN2-59 CNN16-15 +5V CNN16-16 GND CNN16-17 Outer Rotate Switch AC/DC CNN4- CNN16-18 +5V CNN16-19 GND CNN16-20 Inner Rotate Switch AC/DC CNN4- CNN16-21 +5V CNN16-22 GND CNN16-23 Bottle Sensor BTTLSNS AC/DC 7-95 CNN2-95 CNN16-24 +5V CNN16-25 GND CNN16-26 Cup Sensor CUPSNSRO AC/DC 7-96 CNN2-96 CNN16-27 +5V CNN16-28 GND CNN16-29 NC CNN16-30 Cover Open/Close Sensor Open COVER OPEN CNN2-10 CNN16-35+5V CNN16-36 GND CNN16-37 Cover Open/Close Sensor Close COVER CLOSE CNN2-9 CNN16-38 +5V CNN16-39 GND CNN16-40 Stirrer 1 Rotate Home STA1HOME SMC 9-4 CNN3-4 CNN18-10 +5V CNN18-11 GND CNN18-12 Stirrer 1 Rotate Cuvette Sense SENS1 SM AC/DC 8-37 CNN4-37 CNN18-13 +5V CNN18-14 GND CNN18-15 Stirrer 2 Rotate Home STA2HOME SMC 9-5 CNN3-5 CNN18-16 +5V CNN18-17 GND CNN18-18 Stirrer 2 Rotate Cuvette Sense SENS2 SM AC/DC 8-38 CNN4-38 CNN18-19 +5V CNN18-20 GND CNN18-21 Water Bath Full Detect BATHLVL AC/DC 7-58 CNN3-58 CNN18-37 GND CNN18-38 Water Bath Temp Sensor BATHTMP- AC/DC 7-84 CNN3-84 CNN18-39 Water Bath Temp Sensor BATHTMP+ AC/DC 7-83 CNN3-83 CNN18-40 Reagent Barcode Reader TXD2 DAQ 12-11 CNN7-11 CNN20-1 Reagent Barcode Reader RTS2 DAQ 12-13 CNN7-13 CNN20-2 Reagent Barcode Reader RXD2 DAQ 12-12 CNN7-12 CNN20-4 Reagent Barcode Reader CTS2 DAQ 12-14 CNN7-14 CNN20-5 Reagent Barcode Reader TRIGGER2 DAQ 12-15 CNN7-15 CNN20-8 CNN20-9,10 +5V CNN10-11 GND LUI Serial TXD4 DAQ 12-21 CNN7-21 CNN21-1 LUI Serial RTS4 DAQ 12-23 CNN7-23 CNN21-2 LUI Serial RXD4 DAQ 12-22 CNN7-22 CNN21-4 LUI Serial CTS4 DAQ 12-24 CNN7-24 CNN21-5 LUI Serial TRIGGER4 DAQ 12-25 CNN7-25 CNN21-8 CLI Serial Carrier Detect CD5 DAQ 12-26 CNN7-26 CNN22-1 CLI Serial Receive Data RXD5 DAQ 12-27 CNN7-27 CNN22-2 CLI Serial Transmit Data TXD5 DAQ 12-28 CNN7-28 CNN22-3 CLI Serial Data Terminal Ready DTR5 DAQ 12-29 CNN7-29 CNN22-4 CNN22-5 GND CLI Serial Data Set Ready DSR5 DAQ 12-30 CNN7-30 CNN22-6 CLI Serial Request to Send RTS5 DAQ 12-31 CNN7-31 CNN22-7 CLI Serial Clear to Send CTS5 DAQ 12-32 CNN7-32 CNN22-8 CLI Serial Ring Indicator CI5 DAQ 12-33 CNN7-33 CNN22-9 CNN7 CNN22-10 CLI Serial DAQ +5V DAQ Optics Shutter SHUTTER ON DAQ CNN24-1 12-36 CNN7-36 CNN24-2 CNN24-3 +5V CNN24-4 GND Optics Trigger LEDVCC DAQ 12-37 CNN7-37 CNN24-5 Optics Trigger OPT-TRIGGE DAQ 12-39 CNN7-39 CNN24-6 DAQ 12-38 GND CNN24-7 S Pressure Sensor Sig POS SASEN1 POS PM 10-1 11-1 CNN6-1 CNN25-1 +12V CNN25-2 +/-12V CNN25-3 –12V CNN25-4 NC CNN25-5 Pressure Sensor Sig Neg SASEN1 NEG PM 10-2 11-2 CNN6-2 CNN25-6 R1 Pressure Sensor Sig Pos R1ASEN POS PM 10-3 11-3 CNN6-3 CNN25-7 +12V CNN25-8 +/-12V CNN25-9 –12V CNN25-10 NC CNN25-11 R1 Pressure Sensor Sig Neg R1ASEN NEG PM 10-4 11-4 CNN6-4 R2 Pressure Sensor Sig Pos R2ASEN POS PM 10-5 11-5 CNN6-5 R2 Pressure Sensor Sig Neg R2ASEN NEG PM 10-6 11-6 CNN6-6 CNN25-12 CNN25-13 +12V CNN25-14 +/-12V CNN25-15 –12V CNN25-16 NC CNN25-17 CNN25-18 Sample TLLS SLLS CNN6-33 Sample TLLS SLLSG CNN6-34 CNN25-19 CNN25-20 CNN26-1 +5V CNN26-2 +5V Stirrer 1 ON ST1ON AC/DC 7-23 CNN2-23 CNN26-3 Stirrer 1 Scan ST1SCAN AC/DC 7-24 CNN2-24 CNN26-4 Stirrer 1 Clear ST1CLS AC/DC 7-25 CNN2-25 CNN26-5 Stirrer 1 Busy 0 Sig ST1BUSY0 AC/DC 7-26 CNN2-26 CNN26-6 Stirrer 1 Busy 1 Sig ST1BUSY1 AC/DC 7-27 CNN2-27 CNN26-7 Stirrer 2 ON ST2ON AC/DC 7-28 CNN2-28 CNN26-8 Stirrer 2 Scan ST2SCAN AC/DC 7-29 CNN2-29 CNN26-9 Stirrer 2 Clear ST2CLS AC/DC 7-30 CNN2-30 CNN26-10 Stirrer 2 Busy 0 Sig ST2BUSY0 AC/DC 7-31 CNN2-31 CNN26-11 Stirrer 2 Busy 1 Sig ST2BUSY1 AC/DC 7-32 CNN2-32 CNN26-12 CNN2-33 CNN26-16 +5 IN VCC Front Cover Sensor FRONTCVROK AC/DC 7-60 CNN2-60 CNN29-1 +5V CNN29-2 GND CNN29-13 Rear Cover Sensor BACKCVROK AC/DC 7-61 CNN2-61 CNN29-4 +5V CNN29-5 GND CNN29-,6 Rotate Switch ON ISLUILED1(RCLED) AC/DC 7-36 CNN2-36 CNN30-1 +5V CNN30-2 Open/Close Switch ON ISLUISLED2(RCLED) AC/DC 7-38 CNN2-38 CNN30-3 +5V CNN30-4 Rotate Switch ISSW(RCSW) AC/DC 7-91 CNN2-91 CNN30-6 Open/Close Switch ISLSW(RCSW) AC/DC 7-92 CNN2-92 CNN30-7 GND CNN30-8 12V Snsr Si 4 RALOBJECT 4 SM AC/DC 8-32 CNN4-32 CNN31-1 +12V CNN31-2 12VGND CNN31-3 12V Snsr Si RALOBJECT3 SM AC/DC 8-31 CNN4-31 CNN31-4 +12V CNN31-5 12VGND CNN31-,6 12V Snsr Si 2 RALOBJECT3 SM AC/DC 8-30 CNN4-30 CNN31-7 +12V CNN31-8 12VGND CNN31-,9 12V Snsr Si 1 RALOBJECT1 SM AC/DC 8-29 CNN4-29 CNN31-10 +12V CNN31-11 12VGND CNN31-12 Sensor SIG27 SM13DLMT SM AC/DC 8-28 CNN4-28 CNN31-13 +5V CNN31-14 GND CNN31-15 Sensor SIG26 SM13PITCHOK SM AC/DC 8-27 CNN4-27 CNN31-16 +5V CNN31-17 GND CNN31-18 Sensor SIG25 SM13HOME SM AC/DC 8-26 CNN4-26 CNN31-19 +5V CNN31-20 GND CNN31-21 Sensor SIG24 RALOPTION2 SM AC/DC 8-25 CNN4-25 CNN31-22 +5V CNN31-23 GND CNN31-24 Sensor SIG2 RALOPTION1 SM AC/DC 8-24 CNN4-24 CNN31-25 +5V CNN31-26 GND CNN31-27 Sensor SIG22 SM12PITCHOK SM AC/DC 8-23 CNN4-23 CNN31-28 +5V CNN31-29 GND CNN31-30 Sensor SIG21 SM12HOME SM AC/DC 8-22 CNN4-22 CNN31-31 +5V CNN31-32 GND CNN31-33 Sensor SIG20 RALGSNS3 SM AC/DC 8-21 CNN4-21 CNN31-34 +5V CNN31-35 GND CNN31-36 Sensor SIG19 RALGSNS2 SM AC/DC 8-20 CNN4-20 CNN31-37 +5V CNN31-38 GND CNN31-39 Sensor SIG1 RALGSNS1 SM AC/DC 8-19 CNN4-19 CNN31-40 +5V CNN31-41 GND CNN31-42 Sensor SIG17 RALOPEN2 SM AC/DC 8-18 CNN4-18 CNN31-43 +5V CNN31-44 GND CNN31-45 Sensor SIG16 RALOPEN1 SM AC/DC 8-17 CNN4-17 CNN31-46 +5V CNN31-47 GND CNN31-48 Sensor SIG15 AC1UPHOME SM AC/DC 8-33 CNN4-33 CNN31-49 +5V CNN31-50 GND CNN31-51 Sensor SIG14 AC1DOWNMOVE SM AC/DC Sensor SIG13 RALACDOWNOK SM AC/DC Sensor SIG12 SM4HOME SM AC/DC Sensor SIG11 SM4PITCH Sensor SIG10 8-34 CNN4-34 CNN31-52 +5V CNN31-53 GND CNN31-54 CNN2-11 CNN31-55 +5V CNN31-56 GND CNN31-57 8-3 CNN4-3 CNN31-64 +5V CNN31-65 GND CNN31-66 SM AC/DC 8-4 CNN4-4 CNN31-67 +5V CNN31-68 GND CNN31-69 SM4DLMT SM AC/DC 8-5 CNN4-5 CNN31-70 +5V CNN31-71 GND CNN31-72 Sensor SIG9 SM5HOME SM AC/DC 8-6 CNN4-6 CNN31-73 +5V CNN31-74 GND CNN31-75 Sensor SIG8 SM5PITCH SM AC/DC 8-7 CNN4-7 CNN31-76 +5V CNN31-77 GND CNN31-78 Sensor SIG7 SM5DLMT SM AC/DC 8-8 CNN4-8 CNN31-79 +5V CNN31-80 GND CNN31-81 Sensor SIG6 SM6HOME SM AC/DC 8-9 CNN4-9 CNN31-82 +5V CNN31-83 GND CNN31-84 Sensor SIG5 SM6PITCH SM AC/DC 8-10 CNN4-10 CNN31-85 +5V CNN31-86 GND CNN31-87 Sensor SIG4 SM6DLMT SM AC/DC 8-11 CNN4-11 CNN31-88 +5V CNN31-89 GND CNN31-90 Sensor SIG3 SM8HOME SM AC/DC 8-14 CNN4-14 CNN31-91 +5V CNN31-92 GND CNN31-93 Sensor SIG2 SM8PITCH SM AC/DC 8-15 CNN4-15 CNN31-94 +5V CNN31-95 GND CNN31-96 Sensor SIG1 SM8DLMT SM AC/DC 8-16 CNN4-16 CNN31-97 +5V CNN31-98 GND CNN31-99 Sample Line Flow Signal SPFS CNN37- CNN38-1 +5V CNN38-2 GND CNN38-3 Reagent Line Flow Signal RPFS CNN37- CNN38-4 +5V CNN38-5 GND CNN38-6 Stirrer Line Flow Signal STFS CNN37- CNN38-7 +5V CNN38-8 GND CNN38-9 ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Stepper Motor Controller Sensors Description Reaction Crsl Rgt Inner Crsl Rgt Outer Crsl Sample Rotation Sample U/D R1 Arm U/D R1 Arm Rotation R2 Arm U/D R2 Arm Rot Code Card Cage Board Card Cage Back Plane Cable CNN Board CNN Board RUHOME SMC CNN 9- CNN 3-33 CNN18-34 +5V CNN18-35 GND CNN 18-36 RUMIR SMC CNN 9-34 CNN 3-34 CNN18-31 +5V CNN18-32 GND CNN 18-33 R1CIHOME SMC CNN 9-35 CNN 3-35 CNN16-4 +5V CNN16-5 GND CNN 16-6 R1CI PITCH SMC CNN 9-36 CNN 3-36 CNN16-1 +5V CNN16-2 GND CNN 16-3 R1CIPOS SMC CNN 9-37 CNN 3-37 R1COHOME SMC CNN 9-38 CNN 3-38 CNN16-4 +5V CNN16-5 GND CNN 16-12 R1COPITCH SMC CNN 9-39 CNN 3-39 CNN16-7 +5V CNN16-8 GND CNN 16-9 R1COPOS SMC CNN 9-40 CNN 3-40 SAAHOME SMC CNN 9-44 CNN 3-44 CNN15-4 +5V CNN15-5 GND CNN 15-6 SAAMIR SMC CNN 9-45 CNN 3-45 CNN15-1 +5V CNN15-2 GND CNN 15-3 SAEHOME SMC CNN 9-46 CNN 3-46 CNN15-10+5V CNN15-11 GND CNN 15-12 SAEMIR SMC CNN 9-47 CNN 3-47 CNN15-4 +5V CNN15-5 GND CNN 15-9 SACLASH SMC CNN 9-48 CNN 3-48 CNN15-4 +5V CNN15-5 GND CNN 15-16 R1AEHOME SMC CNN 9-17 CNN 3-17 CNN17-10 +5V CNN17-11 GND CNN 17-12 R1AEMIR SMC CNN 9-18 CNN 3-18 CNN17-7 +5V CNN17-8 GND CNN 17-9 R1ACLASH SMC CNN 9-19 CNN 3-19 CNN17-14 +5V CNN17-15 GND CNN 17-16 R1AAHOME SMC CNN 9-15 CNN 3-15 CNN17-4 +5V CNN17-5 GND CNN 17-6 R1AAMIR SMC CNN 9-16 CNN 3-16 CNN17-1 +5V CNN17-2 GND CNN 17-3 R2AEHOME SMC CNN 9-22 CNN 3-22 CNN19-10 +5V CNN19-11 GND CNN 19-12 R2AEMIR SMC CNN 9-23 CNN 3-23 CNN19-7 +5V CNN19-8 GND CNN 19-9 R2ACLASH SMC CNN 9-24 CNN 3-24 CNN19-14 +5V CNN19-15 GND CNN 19-16 R2AAHOME SMC CNN 9-20 CNN 3-20 CNN19-4 +5V CNN19-5 GND CNN 19-6 R2AAMIR SMC CNN 9-21 CNN 3-21 CNN19-1 +5V CNN19-2 GND CNN 19-3 Stirrer 1 STA1HOME SMC CNN 9-4 CNN 3-4 CNN18-10 +5V CNN18-11 GND CNN 18-12 Stirrer 2 STA2HOME SMC CNN 9-5 CNN 3-5 CNN18-16 +5V CNN18-17 GND CNN 18-18 ISE Arm U/D IAEHOME SMC CNN 9-2 CNN 3-2 CNN27-4 +5V CNN27-5 GND CNN 27-6 IAEMIR SMC CNN 9-3 CNN 3-3 CNN27-7 +5V CNN27-8 GND CNN 27-9 ISE Arm Rotation IAAHOME SMC CNN 9-1 CNN 3-1 CNN27-1 +5V CNN27-2 GND CNN 27-3 Sample Pipettor SPHOME SMC CNN 9-6 CNN 3-6 CNN11-4 +5V CNN11-5 GND CNN 11-6 SPMIR SMC CNN 9-7 CNN 3-7 CNN11-1 +5V CNN11-2 GND CNN 11-3 SPDLMT SMC CNN 9-8 CNN 3-8 CNN11-7 +5V CNN11-8 GND CNN 11-9 R1PHOME SMC CNN 9-9 CNN 3-9 CNN11-13 +5V CNN11-14 GND CNN 11-15 R1PMIR SMC CNN 9-10 CNN 3-10 CNN11-10 +5V CNN11-11 GND CNN 11-12 R1PDLMT SMC CNN 9-11 CNN 3-11 CNN11-16 +5V CNN11-17 GND CNN 11-18 R2PHOME SMC CNN 9-12 CNN 3-12 CNN11-22 +5V CNN11-23 GND CNN 11-24 R2PMIR SMC CNN 9-13 CNN 3-13 CNN11-19 +5V CNN11-20 GND CNN 11-21 R2PDLMT SMC CNN 9-14 CNN 3-14 CNN11-25 +5V CNN11-26 GND CNN 11-27 ISOHOME SMC CNN 9-28 CNN 3-28 CNN28-1 +5V CNN28-2 GND CNN 28-3 ISOPITCH SMC CNN 9-29 CNN 3-29 CNN28-4 +5V CNN28-5 GND CNN 28-6 ISOPOS SMC CNN 9-30 CNN 3-30 R1 Pipettor R2 Pipettor HCWP ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Stepper Motor Signal Trace Description Code Card Cage Board Card Cage Back Plane Cable CNN Board Cable CNN Board Cable Driver Box Cable SMD Bd I/F Signal Cable Pins Reaction Carousel Rotate Motor CW RCCW SMC CNN3-51 CNN36-51 CNN33-1 SM1 CNN1-1 Reaction Carousel Rotate Motor CCW RCCCW SMC CNN3-52 CNN36-52 CNN33-2 SM1 CNN1-2 Reagent Inner Carousel Motor CW RICCW SMC CNN3-53 CNN36-53 CNN33-3 SM1 CNN1-3 Reagent Inner Carousel Motor CCW RICCCW SMC CNN3-54 CNN36-54 CNN33-4 SM1 CNN1-4 Reagent Outer Carousel Motor CW ROCCW SMC CNN3-55 CNN36-55 CNN33-5 SM1 CNN1-5 Reagent Outer Carousel Motor CCW ROCCCW SMC CNN3-56 CNN36-56 CNN33-6 SM1 CNN1-6 Sample Arm Up/Down Motor CW SAZCW SMC CNN3-63 CNN36-63 CNN33-7 SM1 CNN1-7 Sample Arm Up/Down Motor CCW SAZCCW SMC CNN3-64 CNN36-64 CNN33-8 SM1 CNN1-8 Sample Arm Rotate Motor CW SAHCW SMC CNN3-61 CNN36-61 CNN33-9 SM1 CNN1-9 Sample Arm Rotate Motor CCW SAHCCW SMC CNN3-62 CNN36-62 CNN33-10 SM1 CNN1-10 R1 Arm Up/Down Motor NA R1ZINA SMC CNN2-66 CNN35-66 CNN33-11 SM1 CNN1-11 R1 Arm Up/Down Motor NB R1ZINB SMC CNN2-67 CNN35-67 CNN33-12 SM1 CNN1-12 R1 Arm Up/Down Motor TDA R1ZTDA SMC CNN2-68 CNN35-68 CNN33-13 SM1 CNN1-13 R1 Arm Up/Down Motor TDB R1ZTDB SMC CNN2-69 CNN35-69 CNN33-14 SM1 CNN1-14 R1 Arm Up/Down Motor Hold R1ZHOLD SMC CNN2-70 CNN35-70 CNN33-15 SM1 CNN1-15 R1 Arm Rotate Motor NA R1HINA SMC CNN2-61 CNN35-61 CNN33-16 SM1 CNN1-16 R1 Arm Rotate Motor NB R1HINB SMC CNN2-62 CNN35-62 CNN33-17 SM1 CNN1-17 R1 Arm Rotate Motor TDA R1HTDA SMC CNN2-63 CNN35-63 CNN33-18 SM1 CNN1-18 R1 Arm Rotate Motor TDB R1HTDB SMC CNN2-64 CNN35-64 CNN33-19 SM1 CNN1-19 R1 Arm Rotate Motor Hold R1HHOLD SMC CNN2-65 CNN35-65 CNN33-20 SM1 CNN1-20 R2 Arm Up/Down Motor NA R2ZINA SMC CNN3-6 CNN36-6 CNN33-21 SM1 CNN1-21 R2 Arm Up/Down Motor NB R2ZINB SMC CNN3-7 CNN36-7 CNN33-22 SM1 CNN1-22 R2 Arm Up/Down Motor TDA R2ZTDA SMC CNN3-8 CNN36-8 CNN33-23 SM1 CNN1-23 R2 Arm Up/Down Motor TDB R2ZTDB SMC CNN3-9 CNN36-9 CNN33-24 SM1 CNN1-24 R2 Arm Up/Down Motor Hold R2ZHOLD SMC CNN3-10 CNN36-10 CNN33-25 SM1 CNN1-25 R2 Arm Rotate Motor NA R2HINA SMC CNN3-1 CNN36-1 CNN33-26 SM1 CNN1-26 R2 Arm Rotate Motor NB R2HINB SMC CNN3-2 CNN36-2 CNN33-27 SM1 CNN1-27 R2 Arm Rotate Motor TDA R2HTDA SMC CNN3-3 CNN36-3 CNN33-28 SM1 CNN1-28 SMD Bd Motor Drive Signal Cable Pins Driver Box Cable CNN2-1 – CNN2-5 SM2,3 CNN4-1 – CNN4-5 SM4,5 CNN5-1 – CNN5-5 CNN3-1 – CNN3-5 SM2,3 CNN6-1 – CNN6-5 SM6 CNN7-1 – CNN7-6 SM7,8 CNN7-7 – CNN7-12 CNN8-1 – CNN8-6 CNN8-7 – CNN8-12 SM8,9 R2 Arm Rotate Motor TDB R2HTDB SMC CNN3-4 CNN36-4 CNN33-29 SM1 CNN1-29 R2 Arm Rotate Motor Hold R2HHOLD SMC CNN3-5 CNN36-5 CNN33-30 SM1 CNN1-30 Stirrer 1 Rotate Motor NA STR1HINA SMC CNN2-11 CNN35-11 CNN33-31 SM1 CNN1-31 Stirrer 1 Rotate Motor NB STR1HINB SMC CNN2-12 CNN35-12 CNN33-32 SM1 CNN1-32 Stirrer 1 Rotate Motor TDA STR1HTDA SMC CNN2-13 CNN35-13 CNN33-33 SM1 CNN1-33 Stirrer 1 Rotate Motor TDB STR1HTDB SMC CNN2-14 CNN35-14 CNN33-34 SM1 CNN1-34 Stirrer 1 Rotate Motor Hold STR1HHOLD SMC CNN2-15 CNN35-15 CN33-35 SM1 CNN1-35 Stirrer 2 Rotate Motor NA STR2HINA SMC CNN2-16 CNN35-16 CNN33-36 SM1 CNN1-36 Stirrer 2 Rotate Motor NB STR2HINB SMC CNN2-17 CNN35-17 CNN33-37 SM1 CNN1-37 Stirrer 2 Rotate Motor TDA STR2HTDA SMC CNN2-18 CNN35-18 CNN33-38 SM1 CNN1-38 Stirrer 2 Rotate Motor TDB STR2HTDB SMC CNN2-19 CNN35-19 CNN33-39 SM1 CNN1-39 Stirrer 2 Rotate Motor Hold STR2HHOLD SMC CNN2-20 CNN35-20 CNN33-40 SM1 CNN1-40 ISE Arm Up/Down Motor NA IAZINA SMC CNN2-6 CNN35-6 CNN33-41 SM1 CNN1-41 ISE Arm Up/Down Motor NB IAZINB SMC CNN2-7 CNN35-7 CNN33-42 SM1 CNN1-42 ISE Arm Up/Down Motor TDA IAZTDA SMC CNN2-8 CNN35-8 CNN33-43 SM1 CNN1-43 ISE Arm Up/Down Motor TDB IAZTDB SMC CNN2-9 CNN35-9 CNN33-44 SM1 CNN1-44 ISE Arm Up/Down Motor Hold IAZHOLD SMC CNN2-10 CNN35-10 CNN33-45 SM1 CNN1-45 ISE Arm Rotate Motor NA IAHINA SMC CNN2-1 CNN35-1 CNN33-46 SM1 CNN1-46 ISE Arm Rotate Motor NB IAHINB SMC CNN2-2 CNN35-2 CNN33-47 SM1 CNN1-47 ISE Arm Rotate Motor TDA IAHTDA SMC CNN2-3 CNN35-3 CNN33-48 SM1 CNN1-48 ISE Arm Rotate Motor TDB IAHTDB SMC CNN2-4 CNN35-4 CNN33-49 SM1 CNN1-49 ISE Arm Rotate Motor Hold IAHHOLD SMC CNN2-5 CNN35-5 CNN33-50 SM1 CNN1-50 Sample Pump Motor SPINA NA SMC CNN2-31 CNN35-31 CNN33-51 SM1 CNN1-51 Sample Pump Motor SPINB NB SMC CNN2-32 CNN35-32 CNN33-52 SM1 CNN1-52 Sample Pump Motor SPTDA TDA SMC CNN2-33 CNN35-33 CNN33-53 SM1 CNN1-53 Sample Pump Motor SPTDB TDB SMC CNN2-34 CNN35-34 CNN33-54 SM1 CNN1-54 Sample Pump Motor SPHOLD Hold SMC CNN2-35 CNN35-35 CNN33-55 SM1 CNN1-55 R1 Pump Motor NA R1PINA SMC CNN2-41 CNN35-41 CNN33-56 SM1 CNN1-56 R1 Pump Motor NB R1PINB SMC CNN2-42 CNN35-42 CNN33-57 SM1 CNN1-57 R1 Pump Motor TDA R1PTDA SMC CNN2-43 CNN35-43 CNN33-58 SM1 CNN1-58 R1 Pump Motor TDB R1PTDB SMC CNN2-44 CNN35-44 CNN33-59 SM1 CNN1-59 R1 Pump Motor Hold R1PHOLD SMC CNN2-45 CNN35-45 CNN33-60 SM1 CNN1-60 R2 Pump Motor NA R2PINA SMC CNN2-51 CNN35-51 CNN33-61 SM1 CNN1-61 R2 Pump Motor NB R2PINB SMC CNN2-52 CNN35-52 CNN33-62 SM1 CNN1-62 CNN9-1 – CNN9-6 CNN9-7 – CNN9-12 CNN10-1 – CNN10-6 SM10,11 CNN10-7 – CNN10-12 CNN11-1 – CNN11-6 CNN11-7 – CNN11-12 CNN12-1 – CNN12-6 SM11,12 R2 Pump Motor TDA R2PTDA SMC CNN2-53 CNN35-53 CNN33-63 SM1 CNN1-63 R2 Pump Motor TDB R2PTDB SMC CNN2-54 CNN35-54 CNN33-64 SM1 CNN1-64 R2 Pump Motor Hold R2PHOLD SMC CNN2-55 CNN35-55 CNN33-65 SM1 CNN1-65 HCW Pump Motor NA PPHCINA SMC CNN3-36 CNN36-36 CNN33-66 SM1 CNN1-66 HCW Pump Motor NB PPHCINB SMC CNN3-37 CNN36-37 CNN33-67 SM1 CNN1-67 HCW Pump Motor TDA PPHCTDA SMC CNN3-38 CNN36-38 CNN33-68 SM1 CNN1-68 HCW Pump Motor TDB PPHCTDB SMC CNN3-39 CNN36-39 CNN33-69 SM1 CNN1-69 HCW Pump Motor Hold PPHCHOLD SMC CNN3-40 CNN36-40 CNN33-70 SM1 CNN1-70 Reagent Handler (1) AUTO1INA Motor NA SMC CNN35-36 CNN33-71 SM1 CNN1-71 Reagent Handler (1) AUTO1INB Motor NB SMC CNN35-37 CNN33-72 SM1 CNN1-72 Reagent Handler (1) AUTO1TDA Motor TDA SMC CNN35-38 CNN33-73 SM1 CNN1-73 Reagent Handler (1) AUTO1TDB Motor TDB SMC CNN35-39 CNN33-74 SM1 CNN1-74 Reagent Handler (1) AUTO1HOLD Motor Hold SMC CNN35-40 CNN33-75 SM1 CNN1-75 Reagent Handler (2) AUTO2INA Motor NA SMC CNN35-46 CNN33-76 SM1 CNN1-76 Reagent Handler (2) AUTO2INB Motor NB SMC CNN35-47 CNN33-77 SM1 CNN1-77 Reagent Handler (2) AUTO2TDA Motor TDA SMC CNN35-48 CNN33-78 SM1 CNN1-78 Reagent Handler (2) AUTO2TDB Motor TDB SMC CNN35-49 CNN33-79 SM1 CNN1-79 Reagent Handler (2) AUTO2HOLD Motor Hold SMC CNN35-50 CNN33-80 SM1 CNN1-80 Reagent Handler (3) AUTO3INA Motor NA SMC CNN35-56 CNN33-81 SM1 CNN1-81 Reagent Handler (3) AUTO3INB Motor NB SMC CNN35-57 CNN33-82 SM1 CNN1-82 Reagent Handler (3) AUTO3TDA Motor TDA SMC CNN35-58 CNN33-83 SM1 CNN1-83 Reagent Handler (3) AUTO3TDB Motor TDB SMC CNN35-59 CNN33-84 SM1 CNN1-84 Reagent Handler (3) AUTO3HOLD Motor Hold SMC CNN35-60 CNN33-85 SM1 CNN1-85 Spare Motor NA AUXINA SMC CNN3-46 CNN36-46 CNN33-86 SM1 CNN1-86 Spare Motor NB AUXINB SMC CNN3-47 CNN36-47 CNN33-87 SM1 CNN1-87 Spare Motor TDA AUXTDA SMC CNN3-48 CNN36-48 CNN33-88 SM1 CNN1-88 Spare Motor TDB AUXTDB SMC CNN3-49 CNN36-49 CNN33-89 SM1 CNN1-89 Spare Motor Hold AUXHOLD SMC CNN3-50 CNN36-50 CNN33-90 SM1 CNN1-90 GND(+5V) SMC CNN35-71 CNN33-91 SM1 CNN1-91 GND(+5V) SMC CNN35-72 CNN33-92 SM1 CNN1-92 GND(+5V) SMC CNN36-71 CNN33-93 SM1 CNN1-93 GND(+5V) SMC CNN36-72 CNN33-94 SM1 CNN1-94 GND(+5V) SMC CNN33-95 SM1 CNN1-95 GND(+5V) SMC CNN33-96 SM1 CNN1-96 +5V SMC CNN35-79 CNN33-97 SM1 CNN1-97 +5V SMC CNN35-80 CNN33-98 SM1 CNN1-98 +5V SMC CNN36-79 CNN33-99 SM1 CNN1-99 +5V SMC CNN36-80 CNN33-100 SM1 CNN1-100 CNN12-7 – CNN12-12 CNN13-1 – CNN13-6 CNN13-7 – CNN13-12 CNN14-1 – CNN14-6 CNN14-7 – CNN14-12 SM13,14 SM14 ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Robotic Sample Handler (RSH) Overview The ARCHITECT c4000 RSH (Robotic Sample Handler) is the interface through which samples are presented to the system for identification and aspiration. The operator places samples into carriers (five sample positions per carrier), which are inserted into the load/unload platform. The Carrier Transport (robotic arm) picks up the sample carrier from the load/unload platform, positions it at the barcode reader for carrier and sample barcode label identification, and places the sample carrier on the Carrier Positioner. The Carrier Positioner moves the sample carrier to the sample aspiration position. After sample aspiration, the Carrier Transport returns the sample carrier to the load/unload platform for removal. Throughout RSH processing, the presence of the carrier is monitored at the following locations: • Load/Unload Platform- a lever mechanism and sensors located underneath each platform detect the presence of a carrier on the load/unload platform. • Carrier Transport Picker Arm- a sensor located on the picker arm detects when a carrier is on the picker arm during transport between the platform, bar code reader or positioner. • Bar Code Reader- an error is generated when the carrier label is not detected. • Carrier Positioner- a reflective sensor on the rear of the Carrier Transport is used to detect the presence of a carrier on the positioner. Hardware Description The RSH contains the following main hardware components: • • • • • • • Load/Unload Platform Load/Unload Sensor Boards Carrier Transport Carrier Positioner Bar Code Reader RSH Distribution Board Card Cage Boards Load/Unload Platform The operator loads sample carriers into the load/unload platform sections, which positions each carrier for transporting by the carrier transport. There are four bays on the load/unload platform of the c4000 RSH. Each bay contains five sections. There are a total of 20 sections separated by partitions. Each section accommodates one sample carrier. In the base of each section is a lever mechanism with a flag, which interacts with a set of sensors mounted on the load/unload sensor board. When a carrier is inserted into a section, the carrier present flag activates the sensors, indicating the presence of the carrier and whether it is fully seated in the section or partially loaded. Load/Unload Sensor Boards There are four load/unload sensor boards that correspond to the four bays. The boards are connected in a daisy chain with a termination connector at the end of the chain. Note The load/unload sensor boards are not compatible with the i1000 load/unload sensor boards. Light emitting diodes (LEDs) attached to the load/unload sensor boards indicate the state of each section to the operator. • • • • • • Indicators off - no sample carrier is loaded in the section Green (steady) - sample carrier is loaded, but processing has not begun. Samples can be accessed Amber (steady) - sample carrier is processing and cannot be accessed Green (blinking) - processing is complete and the sample carrier can be accessed Amber and green (alternating) - bar code scan or other error occurred. Carrier can be accessed Blue- indicates that the section has been configured for priority processing - Up to 7 sections can be configured for priority processing on a stand-alone c4000 RSH - Up to10 sections can be configured for priority processing on a ci4100 integrated system Carrier Transport The carrier transport is a robotic device that interacts with the load/unload platform, bar code reader and the carrier positioner. The carrier transport assembly moves along the length of the load unload platform, riding on a linear rail guide. It interacts with each section to pick up the sample carrier, move it to the bar code reader and then place the carrier on the carrier positioner. It also returns sample carriers from the carrier positioner to the load/unload platform for removal. Carrier Transport Front • • • • Picker Arm: attaches to theta shaft, supports and retains the carrier during pick and place operations Rail Guide Sensor: sensor in picker arm that detects the presence of a sample carrier Theta Motor, Encoder and Shaft: moves carriers in theta axis during pick and place operations Z- Motor, Encoder, Shaft and Sensor: moves carriers in the Z (vertical) direction, raising and lowering carriers during pick and place operations • X- Motor and Encoder: drives the carrier transport along the rail guide in the x axis • Flex Chain Connectors: the flex chain contains two cables that provide the control and power signal information between the carrier transport board and the RSH distribution board. Carrier Transport Rear • Carrier Transport PCB: printed circuit board mounted to the carrier transport frame that interfaces all transport assembly motor and sensor signals to the RSH distribution board via the flex chain cables. • Carrier Positioner Carrier Detect Sensor- senses the presence of a carrier on the Carrier Positioner. • X Motor Gear and Rollers- moves the Carrier Transport horizontally along the rail assembly. 1 Rail 2 Roller 3 X Motor 4 X Motor Gear 5 Gear engages teeth under rail Carrier Positioner The carrier positioner moves the sample carrier into position for sample aspiration by the processing module sample pipettor. The carrier positioner consists of the following major components: • Positioner Tray (contains three pockets)- attached to the slide rail. Each pocket holds one sample carrier • Carrier Positioner Motor- gear moves the positioner tray along the slide rail • Carrier Positioner PCB: interfaces the carrier positioner motor and sensor to the RSH Card Cage. RSH Distribution Board The RSH distribution board is mounted to the left RSH support, behind a protective cover. It provides the interface circuitry between the RSH card cage and the following: • Carrier Transport PCB (via the flex chain)- stepper motors and sensor signals • Load/unload board sensors and LED indicators • Bar code reader signals 1 Flex Chain Connectors at Carrier Transport 2 Bar Code Reader 3 Distribution Board 4 Connector for Load/Unload Boards 5 Flex Chain Connectors at RSH Distribution Board Bar Code Reader The bar code reader mounts to the left support and scans sample carrier ID bar codes and sample ID bar codes. Bar Code Symbologies: • Codabar (checksum is user configurable) • Code 3 of 9 (checksum is user configurable) • Interleaved 2 of 5 (checksum is user configurable) Code 128 subset A, B, and C (checksum is always enabled) Functional Diagram Links RSH Distribution Board 7A_9033a RSH Card Cage The RSH card cage is located on the right side of the c4000 processing module. It contains the Motor Driver Board, Controller Board, Indexer Board and Card Cage Back Plane. Major components of the RSH card cage are: • Card cage fans- forces air downward through the card cage in order to dissipate heat generated by the RSH card cage PCBs • Card Cage Backplane- interface for +5VDC, +12VDC, +36VDC power, analog, and digital signals connecting the RSH Card Cage to the system power supply, SCC, and the electronic devices of the RSH • Controller Board- CPU controller for the RSH • Indexer Board- provides motor control data (speed, direction and power) • Motor Driver Board- provides the drive current to the stepper motor windings. Integrated ci4100 RSH Overview 1 cRSH 2 iRSH 3 SCC The integrated ci4100 RSH consists of the cRSH (c4000 RSH) and iRSH (i1000SR RSH) connected by a long transport rail and platform bridge. Both the cRSH and iRSH are manufactured by Toshiba. The iRSH is manufactured by Toshiba and is not the same as the stand alone i1000SR RSH manufactured by Abbott. The integrated ci4100 RSH receives power and control signals from the c4000 module Main Power Supply and c4000 module RSH Card Cage. The SCC is located inside the i1000SR module. System cRSH iRSH c4000 Toshiba N/A i1000SR N/A Abbott ci4100 Toshiba Toshiba When integrating a new c4000 and a new i1000SR, the i1000SR will arrive without the Abbott RSH. The Toshiba add on iRSH will be installed on the i1000SR module. When integrating a new c4000 with an existing i1000SR, the i1000SR RSH will be removed, discarded and replaced with the Toshiba addon iRSH. Hardware Description 1 RSH Distribution Board There are four bays on the load/unload platform of the cRSH (bays 1-4). Each bay contains five sections. There are a total of 20 sections separated by partitions. Each section accommodates one sample carrier. The cRSH is essentially the same as the stand alone c4000 RSH except for the following: • • • • The bar code reader is moved to the iRSH The transport rail is longer The flex chain is longer The cRSH Load/Unload Sensor Boards are connected in a daisy chain to the iRSH Load/Unload Sensor Boards. The terminator is installed on the last iRSH Load/Unload Sensor Board. Note The cRSH and iRSH Load/Unload Sensor Boards and Load/Unload Platforms are not compatible. 1 Bar Code Reader 2 Bottle Spinner There are four bays on the load/unload platform of the iRSH (bays 5-8). Each bay contains four sections. There are a total of 16 sections separated by partitions. Each section accommodates one sample carrier or reagent carrier. The bar code reader and bottle spinner is located on the iRSH. ci4100 RSH Card Cage and Power Supply The card cage for the ci4100 RSH is located on the right side of the c4000 processing module. It is the same RSH card cage that is used on the stand alone c4000 system. The Main Power Supply in the c4000 module supplies the +5VDC, +12VDC, and+36VDC power for the ci4100 RSH. ci4100 RSH Sample Processing Samples, in primary tubes or sample cups, are placed in sample carriers. The sample carriers are then inserted into sections on the load unload platform. Carriers in the priority sections are processed first, in the order they were inserted, followed by carriers loaded in the routine sections. The user can configure up to ten priority sections on the ci4100 RSH. Once sample carriers are loaded onto the load unload platform, the following occurs: • The carrier transport (CT) moves to the first carrier and picks it up out of the section. The system software and the carrier detection sensors have already determined which sections contain reagent carriers and which contain sample carriers. The system software prioritizes the processing order of the carriers. • The CT moves the carrier to the barcode reader where the carrier ID and sample IDs are scanned. The CT then returns the carrier to its original section. • Based on the SID scanned, the system determines whether a test order request exists for each sample in the carrier. If test orders exist, the CT picks up the carrier and places it on the i1000SR Aspiration Platform or c4000 Carrier Positioner. Module optimization software determines the processing order for each test requested on that carrier. • The pipettor aspirates from each sample, as determined by the optimization (scheduling) software. Once sample aspiration is complete for all samples in the carrier, the CT moves the carrier back to the load unload platform. • Blinking green indicator lights on the load unload platform indicate that sample aspiration for all samples on that carrier has been completed and the carrier can be removed by the operator. ci4100 RSH Reagent Processing The i1000SR reagent kits are loaded onto the i1000SR reagent carousel through the load unload platform in an automated process. Kits are loaded on reagent carriers. Each carrier has a specific, color-coded seat for each type of reagent bottle. The microparticle bottle position is a rotating seat, connected to a cogwheel. This cogwheel engages with the bottle rotator at the bar code reader scan position so that when a reagent carrier is presented for scanning, the microparticle bottle is rotated in front of the bar code reader beam, allowing a successful label read. When a reagent kit is presented to the load unload platform for loading, the system software determines when an acceptable timing opportunity exists in between sample processing, to allow the reagent to be loaded without interrupting the lockstep routine. Upon initiation of the reagent loading, the carrier transport picks up the reagent carrier and moves it to the bar code read position. The CT lowers the carrier until the microparticle seat cogwheel engages with the spinning bottle rotator disk. All 3 bottle positions are then scanned. If a valid read is obtained, the carrier transport then moves the reagent carrier to the reagent carousel load position and loads it on the carousel. The system software automatically schedules an empty reagent kit or a kit with LLS failure for unloading. When there is a timing opportunity as determined by the software, the carrier transport picks up the reagent carrier from the reagent carousel and moves it back to the load unload platform for removal by the operator. Key Diagrams, M&Ds and Alignment/Adjustment Procedures Functional Diagram Links Link RSH Block Diagram 7A_9033a Detailed Top Level 7A_9034a General Top Level 7A_9035a Distribution Board Diagram 7A_9036a Carrier Transport PCB 7A_9037a Carrier Positioner PCB 7A_9038a c4000 Load Platform 7A_9039a ci4100 RSH Extension 7A_9040a Load/Unload Board 7A_9041a Reagent Door and Top Cover Sensors 7A_9045a M&D/Procedure Description M&D 1114 Carrier Transport - Calibrates the Transport with relation to the Load/Unload Platform. - Calibrates the Transport and Positioner relationship. - Calibrates Theta pick and place. M&D 3240 Bar Code Calibration - Calibrates the barcode read position. M&D 1120 Sample Pipettor Calibration - Calibrates the sample probe to the target tool on the carrier positioner. P-657 +36VDC (PS7 Power Supply) Measurement Perform this procedure to measure the +36V on the RSH. P-658 +5VDC (PS8 Power Supply) Measurement Perform this procedure to measure the +5V on the RSH. P-659 +12VDC (PS8 Power Supply) Measurement Perform this procedure to measure the +12V on the RSH. ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. System Control Center (SCC) Overview Operating System and User Interface The System Control Center (SCC) provides a centralized interface through which the operator can control the processing module and RSH via Ethernet connection. The SCC utilizes WINDOWS 2000 operating system. The customer user interface is Abbott proprietary software. • Base Configuration- loads the Windows 2000 operating system. • Application software- loads the customer user interface. SCC Technical Information Refer to the ARCHITECT c4000 Operations Manual for information regarding: • Software interface • Snapshot screen • Log on • Screens, windows, messages, prompts and refresh buttons • Software navigation Refer to the c4000 Instrument Service Advisories (ISAs) for the latest information regarding: • SCC Platform types • Unpacking instructions • Cable connections. Refer to the ARCHITECT c4000 Service & Support Manual for information regarding: • Monitor support arm installation • Touch screen monitor installation • SCC hardware and cabling installation • SCC installation verification Refer to the c4000 Technical Service Bulletins (TSBs) for the latest information regarding: • • • • • • Base Configuration software and installation instructions Application software and installation instructions (upgrades and new system installs) Module configuration instructions Verification and editing of SCC BIOS Parameters SCC Disaster recovery SCC Intel Active Monitoring SCC Interface Diagrams Functional Diagram Links Link c4000 SCC Interface Diagram 7A_9042a ci4100 SCC Interface Diagram 7A_9043a ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Card Cage Links AC/DC Controller CPU DAQ Main Card Cage MECH LED New Style MECH LED PM SMC SM AC/DC ARCHITECT c4000 Service and Support Manual (Version 204733-122) • © 2009 • 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. AC/DC Controller The AC/DC-Cont-C (controller) board controls AC motors and magnetic DC valves. The board provides monitoring (at test point), and adjustment (via variable resistors) for the temperature sensors. Power supply voltages (VDC) are monitored: • • • • • +5 ±15 +12 +11.5V +24VDC AC/DC Controller Board (side view) S1 & S2 Switch Settings Switch Function Setting S1-1 UNUSED OFF S1-2 UNUSED OFF S1-3 UNUSED OFF S1-4 UNUSED OFF S1-5 UNUSED OFF S1-6 UNUSED OFF S1-7 UNUSED OFF S1-8 UNUSED OFF S2-1 New Style CPU with Flash Memory ON S2-2 UNUSED OFF S2-3 UNUSED OFF S2-4 UNUSED OFF S2-5 UNUSED OFF S2-6 UNUSED OFF S2-7 UNUSED OFF S2-8 c4000 ON AC/DC Controller Board (end view) AC/DC Block Diagram 1 Control EPLD The control EPLD (Electrically Programmable Logic Device) includes the drive circuits for the AC motors and DC valves; contains a majority logic circuit for the sensor input, a sensor and status input register; and interfaces these circuits to the bus. 2 Temperature measurement circuit Temperature information form the temperature sensors is input to this circuit (8-bit A/D converter. This circuit supports five temperature sensors: for the water bath, for the reagent supply center 1, for the reagent supply center 2, for the sample carousel, and the internal temperature sensor. (See Temperature Sensor Adjustment.) 3 Power Supply Voltage Monitoring Circuit This circuit monitors six voltage levels to allow the host CPU to determine whether each power supply has reached the minimum specified voltage. (See Power Supply Monitored Detection Voltages.) 4 Other Output Circuits In addition to the circuits mentioned above, the AC/DC Controller board includes the mixer driver interface circuit, LED drive circuit, and buzzer. 5 General Purpose DIPSW This is a general purpose DIPSW input port used during development. (all switches should be in the down position) 6 General Purpose LED This is a general purpose LED display port used during development. 7 Reset Circuit The reset circuit outputs the reset signal to each circuit when the power is turned ON or the reset SW is activated. 8 PCI Interface Circuit This circuit controls the interface bus with the host CPU. This circuit uses an EPLD. 9 Power ON LED This green LED lights when the+5VDC power is supplied. Temperature Sensor Adjustment VR No. Temperature Sensor Code VR1 Reagent Supply Center No. 1 R1CTEMP VR2 Reagent Supply Center No. 2 R2CTEMP VR3 Sample Carousel ISTEMP VR4 Internal Temperature IVRSTR VR5 Water Bath Temperature BATHTEMP Temperature Sensor Adjustment Test Pins TP No Test Point TP No Test Point TP3 R1 Reagent Temp Sensor (R1CTEMP) Input Voltage TP1 R1CTEMP, R2CTEMP REF Voltage (5V) TP4 R2 Reagent Temp Sensor (R2CTEMP) Input Voltage TP2 R1CTEMP, R2CTEMP VF/2 Input (2.5V) TP7 Sample Carousel Temp Sensor (ISTEMP) Input Sensor TP5 ISTEMP, IVRSTR REF Voltage (5V) TP8 Water Bath Cooling Temp Sensor (IVRSTR) Input Voltage TP6 ISTEMP, IVRSTR VR/2 Input (2.5V) TP11 Bath Temp Sensor (BATHTEMP) Input Voltage TP9 BATHTEMP REF Voltage Output (5V) TP25 GND TP10 BATHTEMP REF VF/2 Input (2.5V) Power Supply Monitored Detection Voltages Monitoring voltage Detection voltage +5 VDC +4.75 to +4.85 (V) +15 VDC +13.58 to +14.42 (V) -15 VDC -13.58 to -14.42 (V) +12 VDC +10.77 to +11.43 (V) +11.5 VDC +10.49 to +11.12 (V) +24 VDC +21.84 to +23.18 (V) AC/DC Controller PCI Bus Interface Connectors (J1/J2) 22 GA4 GA3 GA2 GA2 GA0 GND 21 CLK6 GND RSV RSV RSV GND 20 CLK5 GND RSV GND RSV GND 19 GND GND RSV RSV RSV GND 18 BRSVP2A18 BRSVP2B18 BRSVP2C18 GND BRSVP2E18 GND 17 BRSVP2A17 GND PRST# REQ6# GNT6# GND 16 BRSVP2A16 BRSVP2B16 DEG# GND BRSVP2E16 GND 15 BRSVP2A15 GND FAL# REQ5# GNT5# GND 14 AD[35] AD[34] AD[33] GND AD[32] GND 13 AD[38] GND V(I/O) AD[37] AD[36] GND 12 AD[42] AD[41] AD[40] GND AD[39] GND 11 AD[45] GND V(I/O) AD[44] AD[43] GND 10 AD[49] AD[48] AD[47] GND AD[46] GND 9 AD[52] GND V(I/O) AD[51] AD[50] GND 8 AD[56] AD[55] AD[54] GND AD[53] GND 7 AD[59] GND V(I/O) AD[58] AD[57] GND 6 AD[63] AD[62] AD[61] GND AD[60] GND 5 C/BE[5]# GND V(I/O) C/BE[4]# PAR64 GND 4 V(I/O) BRSVP2B4 C/BE[7]# GND C/BE[6]# GND 3 CLK4 GND GNT3# REQ4# GNT4# GND 2 CLK2 CLK3 SYSEN# GNT2# REQ3# GND 1 CLK1 GND REQ1# GNT1# REQ2# GND 25 5V REQ64# ENUM# 3.3V 5V GND 24 AD[1] 5V V(I/O) AD[0] ACK64# GND 23 3.3V AD[4] AD[3] 5V AD[2] GND 22 AD[7] GND 3.3V AD[6] AD[5] GND 21 3.3V AD[9] AD[8] M66EN C/BE[0]# GND 20 AD[12] GND V(I/O) AD[11] AD[10] GND 19 3.3V AD[15] AD[14] GND AD[13] GND 18 SERR# GND 3.3V PAR C/BE[1]# GND 17 3.3V SDONE SBO# GND PERR# GND 16 DEVSEL# GND V(I/O) STOP# LOCK# GND 15 3.3V FRAME# IRDY# GND TRDY# GND 14 KEY AREA 13 12 11 AD[[18] AD[17] AD[16] GND C/BE[2]# GND 10 AD[21] GND 3.3V AD[20] AD[19] GND 9 C/BE[3]# IDSEL AD[23] GND AD[22] GND 8 AD[26] GND V(I/O) AD[25] AD[24] GND 7 AD[30] AD[29] AD[28] GND AD[27] GND J2 Connector J1 Connector 6 REQ# GND 3.3V CLK AD[31] GND 5 BRSVP1A5 BRSVP1B5 RST# GND GNT# GND 4 BRSVP1A4 GND V(I/O) INTP INTS GND 3 INTA# INTB# INTC# 5V INTD# GND 2 TCK 5V TMS TDO TDI GND 1 5V -12V TRST# +12V 5V GND Pin No A B C D E F Motor Driver Signals and Sensors Pin Assignments (J4/J5) A and B Pin Signal Name Symbol A22 +24V A21 Value I/O Pin Signal Name Symbol Value I/O +24V I B22 +24V +24V I +12V +12V I B21 +12V +12V I A20 +11.5V +11.5V I B20 +11.5V +11.5V I A19 +15V +15V I B19 +15V +15V I A18 +15V +15V I B18 +15V +15V I A17 AGND AGND I B17 AGND AGND I A16 AGND AGND I B16 AGND AGND I A15 Water Bath Cooling Temp Sensor (-) IVRSTR- AI B15 Sample Carousel Temp Sensor (-) ISTEMP- AI A14 Water Bath Cooling Temp Sensor (+) IVRSTR+ AI B14 Sample Carousel Temp Sensor (+) ISTEMP+ AI A13 GND GND I B13 GND GND I A12 GND GND I B12 GND GND I A11 GND GND I B11 GND GND I A10 +5V +5V I B10 +5V +5V I A09 +5V +5V I B09 +5V +5V I A08 Cell Wash CW Signal WSMUP L O B08 AC Motor Spare1 CW Signal AUX1UP A07 Cell Wash CCW Signal WSMDOWN L O B07 A06 Cell Wash Brake Signal WSMBK L O A05 Stirrer Up/Down CW Signal STEMUP L A04 Stirrer Up/Down CCW Signal STEMDOWN A03 Stirrer Up/Down Brake Signal A02 L O AC Motor Spare1 AUX1DOWN CCW Signal L O B06 AC Motor Spare1 Brake Signal AUX1BK L O O B05 Wash Pump Rotate Signal DPMUD L O L O B04 Wash Pump Brake Signal DPMBK L O STEMBK L O B03 Cell Wash Pump Rotate Signal CWPMUD L O Probe Wash Pump CW Signal PWPMUP L O B02 Cell Wash Pump Brake Signal CWPMBK L O A01 Probe Wash Pump CCW Signal PWPMDOWN L O B01 HCW Pump Rotate Signal TWPMUD L O A25 Probe Wash Pump Brake Signal PWPMBK L O B25 HCW Pump Brake Signal TWPMBK L O B24 B3 Empty Sensor A24 J5 Connector J4 Connector (SS) A23 Outside PS (+12V) DALARMV I B23 B14 Empty Sensor (SS) A22 Degasser Vac Tank Error DALARM I B22 MCBC Empty Sensor (SS) A21 B21 Spare Sensor A20 B20 B3 Empty Sensor L A19 B14 Empty Sensor ESB14 H I B19 IS-LUI Push Switch ISSW L I A18 Wash Unit Level Sensor 5 FFS5 L I B18 Sample Carousel Cover Sensor ISCVROK L I A17 Wash Unit Level Sensor 2 FFS2 L I B17 Wash Unit Water Level Sensor 3 FFS3 L I A16 R2 Supply Cover Sensor R2CCVROK L I B16 R1 Supply Cover Sensor R1CCVROK L I A15 Spare DC Motor UP SENSOR AUXMUP H I B15 Cuvette Water Level Sensor BATHLVL L I 14 KEY AREA 13 12 A11 AC Motor Spare 2 UP SENSOR AC2UP H I B11 AC Motor Spare 2 Down Sensor AC2DOWN H I A10 AC Motor Spare 1 DOWN SENSOR AC1DOWN H I B10 AC Motor Spare 1 DOWN OK SENSOR AC1OK H I A09 ISE Ref Pump UP SENSOR IMPUP H I B09 AC Motor Spare 1 UP SENSOR AC1UP H I A08 ISE Ref Pump UP SENSOR ISPUP H I B08 ISE Ref Pump DOWN SENSOR ISPDOWN H I A07 HCW Pump UP SENSOR TWPUP H I B07 HCW Pump DOWN SENSOR TWPDOWN H I A06 Probe Wash Pump OK Sensor PWPUP H I B06 Probe Wash Pump DOWN SENSOR PWPDOWN H I A05 Cell Wash Pump OK Sensor CWPUP H I B05 Cell Wash Pump DOWN SENSOR CWPDOWN H I A04 Wash Pump UP SENSOR DPUP H I B04 Wash Pump DOWN SENSOR DPDOWN H I A03 Stirrer UP/DOWN SENSOR STEDOWN H I B03 Stirrer DOWN OK SENSOR STEDOK H I A02 Cell Wash DOWN OK SENSOR WSDOWNOK H I B02 Stirrer Up/Down OK Sensor STEUP H I A01 Cell Wash UP SENSOR WSUP H I B01 Cell Wash DOWN SENSOR WSDOWN H I Motor Driver Signals and Sensors Pin Assignments (J4/J5) C and D Pin Signal Name Symbol Value I/O Pin Signal Name Symbol Value I/O C22 D22 GND (+24V) GND(+24V) I C21 D21 GND (+12V) GND(+12V) I C20 D20 GND (+11.5V) GND(+11.5V) I C19 D19 -15V -15V I C18 D18 -15V -15V I C17 D17 AGND AGND I C16 D16 AGND AGND I C15 R2 Supply Temp Sensor (-) R2CTEMP- AI D15 R1 Supply Temp Sensor (-) R1CTEMP- AI C14 R2 Supply Temp Sensor (+) R2CTEMP+ AI D14 R1 Supply Temp Sensor (+) R1CTEMP+ Ai C13 GND GND I D13 GND GND I C12 GND GND I D12 GND GND I C11 GND GND I D11 GND GND I C10 +5V +5V I D10 +5V +5V I C09 +5V +5V I D09 +5V +5V I C08 ISE Ref Pump Rotate Signal ISPMUD L O D08 Power Save Relay1 ON Signal MC1ON L O C07 ISE Ref Pump Brake Signal ISPMBK L O D07 Power Save Relay2 ON Signal MC2ON L O C06 ISE Ref Pump Rotate Signal IMPMUD L O D06 AC Motor Spare 3 ON Signal AUX3ON L O C05 ISE Ref Pump Brake Signal IMPMBK L O D05 ISE Asp Pump Valve ISPVON L O C04 AC Motor Spare 2 Rotate Signal AUX2UD L O D04 DI Valve 1 FV1ON L O C03 AC Motor Spare 2 Brake Signal AUX2BK L O D03 DI Valve 2 FV2ON L O C02 Cell Wash Vac Pump ON Signal CWPMON L O D02 DI Valve 3 FV3ON L O C01 Cell Dryer Vac Pump ON Signal CDVPMON L O D01 DI Valve 4 FV4ON L O C25 DI Supply Pump 1ON Signal FP1ON L O D25 DI Valve 5 FV5ON L O C24 DI Supply Pump 2ON Signal FP2ON L O D24 C23 D23 C22 ISLSW Switch ISLSW L I D22 C21 Spare Sensor 2 AUXSW2 L I D21 C20 Sample Cup Sensor C/C CUPSNSRO L I D20 C19 MCC Empty Sensor ESMCC H I D19 C18 Drain Tank Float SW DTF L I D18 C17 Wash Unit Level Sensor 4 FFS4 L I D17 C16 Wash Unit Level Sensor FFS1 L I D16 Water Bath Heater ON Signal IHTRON L O C15 Spare DC Motor Down Sensor AUXMDOWN H I D15 Water Bath Pump ON Signal IPMON L O 14 13 KEY AREA 12 C11 Status Indicator LED 14 CPULED14 L O D11 Status Indicator LED 15 CPULED15 L O C10 Status Indicator LED 12 CPULED12 L O D10 Status Indicator LED 13 CPULED13 L O C09 Status Indicator LED 10 CPULED10 L O D09 Status Indicator LED 11 CPULED11 L O C08 Status Indicator LED 8 CPULED8 L O D08 Status Indicator LED 9 CPULED9 L O C07 Status Indicator LED 6 CPULED6 L O D07 Status Indicator LED 7 CPULED7 L O C06 Status Indicator LED 4 CPULED4 L O D06 Status Indicator LED 5 CPULED5 L O C05 Status Indicator LED 2 CPULED2 L O D05 Status Indicator LED 3 CPULED3 L O C04 Status Indicator LED 0 CPULED0 L O D04 Status Indicator LED 1 CPULED1 L O C03 Spare LED 1 AUXLED1 L O D03 Spare LED 2 AUXLED2 L C02 IS-LUI LED1 ISLUILED1 L O D02 IS-LUI LED2 ISLUILED2 L O C01 Disk Sampler LED1 ISLED1 L O D01 Disk Sampler LED2 ISLED2 L O Value I/O Motor Driver Signals and Sensors Pin Assignments (J4/J5) E and F Pin Signal Name Symbol E22 GND (+24V) E21 Value I/O Pin Signal Name GND(+24V) I F22 GND GND (+12V) GND(+12V) I F21 GND E20 GND (+11.5V) GND(+11.5V) I F20 GND E19 -15V -15V I F19 GND E18 -15V -15V I F18 GND E17 AGND AGND I F17 GND E16 AGND AGND I F16 GND E15 Cuvette Temp Sensor (-) BATHTMP- AI F15 GND E14 Cuvette Temp Sensor (+) BATHTMP+ F14 GND E13 GND GND I F13 GND E12 GND GND I F12 GND E11 GND GND I F11 GND E10 +5V +5V I F10 GND E09 +5V +5V I F09 GND E08 Cell Wash Valve ON Signal CWON L O F08 GND E07 Alkaline Wash Valve ON Signal DTVB14ON L O F07 GND E06 Acid Wash Valve ON Signal DTVB3ON L O F06 GND E05 Vac Pump Valve ON Signal DRYVON L O F05 GND E04 Sampling Pump Valve ON Signal SPVON L O F04 GND E03 Reagent Pump1 Valve ON Signal RP1VON L O F03 GND E02 Reagent Pump2 Valve ON RP2VON L O F02 GND Symbol Signal E01 Dryer Probe Individual ON Signal WSSOLEON L O F01 GND E25 DC Value Spare Valve ON Signal AUXVON L O F25 GND E24 DC Motor Spare Out CW Signal AUXMCW L O F24 GND E23 DC Motor Spare Out CCW Signal AUXMCCW L O F23 GND F22 GND F21 GND E20 F20 GND E19 F19 GND E18 F18 GND E22 E21 Buzzer ON Signal BUZZER L O E17 Water Bath Cooling Fan ON Signal IFANON L O F17 GND E16 Water Bath Waste Valve ON Signal IMVOPEN L O F16 GND E15 Water Bath Waste Valve Close ON Signal IMVSHUT L O F15 GND I F11 GND 14 KEY AREA 13 12 E11 External +5V +5V (IN) E10 Stirrer Cont ST2BUSY1Signal ST2BUSY1 L I F10 GND E09 Stirrer Cont ST2BUSYOSignal ST2BUSY0 L I F09 GND E08 Stirrer Cont ST1BUSY1Signal ST1BUSY1 L I F08 GND E07 Stirrer Cont ST1BUSY0Signal ST1BUSY0 L I F07 GND E06 Stirrer Cont ST2CLRSignal ST2CLR L O F06 GND E05 Stirrer Cont ST2SCANSignal ST2SCAN L O F05 GND E04 Stirrer Cont ST2ONSignal ST2ON L O F04 GND E03 Stirrer Cont ST1CLRSignal ST1CLR L O F03 GND E02 Stirrer Cont ST1SCANSignal ST1SCAN L O F02 GND E01 Stirrer Cont ST1ONSignale ST1ON L O F01 GND Front Panel Interface Connector (CNN1) Connector Pin Map Pin No Signal Name Symbol Value I/O Pin No Signal Name 1 Cell Wash UP SENSOR WSUP L O 26 2 Cell Wash DOWN SENSOR WSDOWN L O 3 Cell Wash DOWN OK SENSOR WSDOWNOK L 4 Stirrer Up/Down UP SENSOR STEUP 5 Stirrer Up/Down DOWN SENSOR 6 Stirrer Up/Down DOWN Symbol Value I/O R1 Supply Cover Sensor R1CCVROK L O 27 R2 Supply Cover Sensor R2CCVROK L O O 28 Wash Unit Level Sensor 1 FFS1 L O L O 29 Wash Unit Level Sensor 2 FFS2 L O STEDOWN L O 30 Wash Unit Level Sensor 3 FFS3 L O STEDOK L O 31 Wash Unit Level FFS4 L O OK SENSOR Sensor 4 7 Wash Pump UP SENSOR DPUP L O 32 Wash Unit Level Sensor 5 FFS5 L O 8 Wash Pump DOWN SENSOR DPDOWN L O 33 Round Cover Sampler Sensor ISCVROK L O 9 Cell Wash Pump UP SENSOR CWPUP L O 34 Drain Tank Float SW DTF L O 10 Cell Wash Pump DOWN SENSOR CWPDOWN L O 35 Degasser Vac Tank Error DALARM L O 11 Probe Wash Pump UP SENSOR PWPUP L O 36 IS-LUI Push Switch ISSW L O 12 Probe Wash Pump DOWN SENSOR PWPDOWN L O 37 MCC Empty Sensor ESMCC L O 13 HCW Pump UP SENSOR TWPUP L O 38 B14 Empty Sensor ESB14 L O 14 HCW Pump DOWN SENSOR TWPDOWN L O 39 B3 Empty Sensor ESB3 L O 15 ISE ASP Pump UP SENSOR ISPUP L O 40 Sample Cup Sensor C/C CUPSNSR1 L O 16 ISE ASP Pump DOWN SENSOR ISPDOWN L O 41 Sample Cup Sensor STAT CUPSNSR0 L O 17 ISE Ref Pump UP SENSOR IMPUP L O 42 Spare Sensor 1 AUXSW1 L O 18 Spare 1 UP SENSOR AC1UP L O 43 Spare Sensor 2 AUXSW2 L O 19 Spare 1 DOWN SENSOR AC1DOWN L O 44 MCC Empty Sensor (SS) ESMCCSS L O 20 Spare 2 DOWN OK SENSOR AC1OK L O 45 B14 Empty Sensor (SS) ESB14SS L O 21 Spare 2 UP SENSOR AC2UP L O 46 B3 Empty Sensor (SS) ESB3SS L O 22 Spare 2 DOWN SENSOR AC2DOWN L O 47 Power +5V O 23 DC Motor Spare UP SENSOR AUXMUP L O 48 Power +5V O 24 DC Motor Spare DOWN SENSOR AUXMDOWN L O 49 Power +5V O 25 Cuvette Water Level Sensor BATHLVL` L O 50 Power +5V O Connector CNN2 Pin Map (Connector is located at the front side of the connector board and provides the signal for multi purpose LED) Pin No Signal Name Symbol Value I/O 1 Multipurpose LED 00bit LEDOUT00 L O 2 Multipurpose LED 01bit LEDOUT10 L O 3 Multipurpose LED 02bit LEDOUT20 L O 4 Multipurpose LED 03bit LEDOUT30 L O 5 Power +5V O 6 Power GND O AC/DC CN1 Connector (True Value = Low) Pin No. Signal Name Code Pin No. Signal Name Code 1 Cuvette washer UP sensor WSUP 26 Reagent supply center 1 cover sensor R1CCVROK 2 Cuvette washer DOWN sensor WSDOWN 27 Reagent supply center 2 cover sensor R2CCVROK 3 Cuvette washer DOWN OK sensor WSDOWNOK 28 4 Mixer elevator UP sensor STEUP 29 Water tank upper limit float switch FFS2 5 Mixer elevator DOWN sensor STEDOWN 30 Water tank level high float switch FFS3 6 Mixer elevator DOWN OK sensor STEDOK 31 Water tank level low float switch FFS4 7 Wash solution pump UP sensor DPUP 32 8 Wash solution pump DOWN sensor DPDOWN 33 Sample carousel cover switch ISCVROK 9 Cuvette wash pump UP sensor CWPUP 34 Low Concentration waste tank level high float switch DTF 10 Cuvette wash pump DOWN sensor CWPDOWN 35 Degasser vacuum tank abnormal sensor DALARM 11 Inner probe wash pump UP sensor PWPUP 36 IS LUI ISSW 12 Inner probe wash pump DOWN sensor PWPDOWN 37 ICT reference solution empty sensor ESMCC 13 H.C. waste pump UP sensor TWPUP 38 Alkaline bulk solution empty sensor ESB14 14 H.C. waste pump DOWN sensor TWPDOWN 39 Acid bulk solution empty sensor ESB3 15 ICT aspiration pump UP sensor ISPUP 40 C/C CUPSNSR1 16 ICT aspiration pump DOWN sensor ISPDOWN 41 STAT CUPSNSR0 17 ICT reference solution pump UP sensor IMPUP 42 Sample carousel pause switch AUXSW1 18 AC1UP 43 Sample carousel scan switch AUXSW2 19 AC1DOWN 44 MCC (SS) ESMCCSS 20 AC11OK 45 B14 (SS) ESB14SS 21 AC2UP 46 B3 (SS) ESB3SS 22 AC2DOWN 47 Power supply +5VDC 23 AUXMUP 48 Power supply +5VDC 24 AUXMDOWN 49 Power supply +5VDC BATHLVL 50 Power supply +5VDC 25 Water bath level sensor FFS1 FFS5 ARCHITECT c4000 Service and Support Manual (Version 204733-112) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. CPU CPU Board Component Description 1 Compact Flash Memory 2 INTEL Celeron 566MHz CPU 3 Phoenix BIOS ROM CPU Board Specifications Note Processor Memory Interface When installing a CPU Board with flash memory [1], verify the AC/DC Controller Board is jumpered correctly. S2 switch one (1), located on the AC/DC Controller Board, is in the ON position for a CPU with Flash Memory and OFF for a CPU with Disk On Chip. CPU [2] Intel Celeron 566MHz CHIPSET Intel 815E Secondary cache Pipeline burst SRAM 256 KB (512 MB maximum) System memory 3.3 V SDRAM DIMM 32 MB (256 MB maximum) BIOS ROM Phoenix BIOS [3] IDE interface Equipped with PCI ICE (1 channel), PIO modes 0 to 4 supported. Capable of supporting Ultra DMA mode. Floppy interface Equipped with a 3.5" FDD port (1 channel) Serial Port (COM 1 & COM 2) Equipped with 16550 compatible port (2 channels, RS-232C compliant. Parallel Port (LPT) Equipped with a standard parallel port (1 channel) two way data transfer supported. Keyboard interface Equipped with a PS/2 keyboard Front Panel Environment CPU Front Connectors CPU Block Diagram Ethernet interface Equipped with 10BASE-T or 100BASE-TX Compact PCI bus interface Address/data bus: 32bit Bus clock: 33 MHz V (I/O): +5 VDC Video interface (optional) Video RAM 2 MB Display mode640 x 480, 16 colors/ 256 colors 1024 x 768, 256 colors/ 64 K colors Switch Generates NMI or reset signals LED PWR (green) ON = +5 VDC HDD (green) ON = HDD is accessed LINK (green) ON = Ethernet link is enabled ACT (yellow) ON = Ethernet transmit or receive Line voltage +5 VDC ± 5% Current consumption 5 amps maximum (5 VDC) Ambient temperature During operation = 0°C to 50° (forced air cooling with an air velocity of 3 m/s must be provided) During non-operation = -20°C to 70°C Ambient humidity During operation = 10% to 90% (no condensing) CPU Front Cover Connectors 1 Ethernet Cable from CPU 2 D-sub connector for CLI 3 High concentration waste tank connector 4 RSH ethernet cable Connector KB (Keyboard connector (PS/2 style keyboard) COM1 & COM2 (RS-232C serial ports) LPT (Standard parallel port connector) Pin # Signal Name 1 DATA 2 RESERVED 3 GND 4 +5VDC 5 CLOCK 6 RESERVED 1 Data Carrier Detect 2 Receive Data 3 Transmit Data 4 Data Terminal Ready 5 GND 6 Data Set Ready 7 Request to Send 8 Clear to Send 9 Ring Indicate 1 STROBE 2 DATA BIT 0 3 DATA BIT 1 4 DATA BIT 2 5 DATA BIT 3 6 DATA BIT 4 7 DATA BIT 5 8 DATA BIT 6 9 DATA BIT 7 10 ACK 11 BUSY 12 PE 13 SLCT 14 AUTO FD TX 15 ERROR 16 INIT 17 SLCT IN 18 GND 19 GND 20 GND 21 GND 22 GND 23 GND Reference Video (optional - Video connector supports VGA or SVGA when video card present) ETHERNET (Ethernet connector supports CAT5 100 BASE-TX/10BASE-T) 24 GND 25 GND 1 RED 2 GREEN 3 BLUE 4 NOT CONNECTED 5 AGND 6 AGND 7 AGND 8 AGND 9 NOT CONNECTED 10 AGND 11 NOT CONNECTED 12 DDC_SDA 13 HSYNC 14 VSYNC 15 DDC_SCL 1 TD+ 2 TD- 3 RD+ 4 RESERVED 5 RESERVED 6 RD- 7 RESERVED 8 RESERVED NMI RESET (This switch generates a NMI to restart the system) PWR (This LED is on when the power is ON) LINK (This LED is on when the system is linked to an Ethernet) HDD (This LED is on when the optional hard drive is accessed) ACT (This LED is on during Ethernet transmissions or receptions) ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. DAQ The DAQ board collects optical and ICT (ISE) measurement data. The board incorporates the following functions and capabilities: • • • • • • • • 386EX Intel CPU 256 Kbytes SRAM for the 386EX Intel CPU 16 bit A/D Converter function 256 Kbytes SRAM for data exchange between 386EX Intel CPU and the host CPU Transfers data from the A/D converter to the host CPU through distraction present (DP) memory Monitors the +5VDC circuit - indicated by the green LED Flash memory (256 Kbytes) for application, boot, and diagnostic programs for the 386EX Intel CPU SRAM backup circuit lasts for 2 weeks DAQ Board DAQ Board Front Panel CN1 Connector for recorder signals for electrodes (Na, K, and Cl). Signal Assignment on CN1 S3, S4, & S5 Rotary switches used for channel selection, must be enabled by W2, W3 & W4. Switches should be set to 0. CN2 Debugging serial interface connector. The SIO built in the 386EX is connected to this connector. DS1 Debugging LED. PIOs (0 to 3 of port 1) built in the 386EX are connected to these LEDs. DS2 Power LED on when +5 VDC is supplied to this board. S2 Debugging/boot selection switch. Refer to tables below for S2 switch default settings. S6 Status can be read by the register on the PCI bus. Set rotary switch to 1. S1 Reset switch for 386EX on this board. DAQ Bd 32 Switch Settings for P/N 7-93181-04 and lower revisions Switch Setting S2-1 OFF S2-2 OFF S2-3 OFF S2-4 OFF S2-5 ON S2-6 ON DAQ Bd 32 Switch Settings for P/N 7-93181-05 and higher revisions Switch Setting S2-1 OFF S2-2 OFF S2-3 OFF S2-4 OFF S2-5 OFF S2-6 OFF DAQ Block Diagram 1 386EX Transfers A/D converted data from analog circuit section to host CPU via DP RAM. 2 Voltage Monitoring Circuit Monitors +5 volts DC and outputs reset signals to each circuit. 3 Control Circuit (EPLD) Indicates EPLD (Electrically Programmable Logic Device) and controls the 386EX CPU. 4 Flash Memory 1 For application programs and data for 386EX CPU. Memory size: 256 KB. 5 Flash Memory 2 For for boot and test programs for the 386EX CPU. Memory size: 256 KB. 6 SRAM Used by 386EX CPU. Memory size: 256 KB. 7 16-bit A/D Conversion Analog Section 16 channel analog input converted from analog to 16 bit digital. 8 PCI Target Interface Circuit. Slave interface circuit for compact PCI bus. 9 Arbitration Control (EPLD) Includes an EPLD and arbitrates access conflicts between 386EX CPU and host CPU. 10 SRAM Used for data exchange between 386EX CPU and host CPU. Memory size is 256 KB. Circuit requires battery backup. 11 Backup Circuit Backup circuit for SRAM (10). The backup life is two weeks. 12 SIO Serial port accessed by the host CPU via the PCI bus. Has six channels. 13 LED and DIP Switch DIP switch and LED display are for use in development. 14 Power ON LED Power ON when +5 Volts DC power is supplied. ICT Signals Optics ARCHITECT c4000 Service and Support Manual (Version 204733-116) • © 2009, 2012 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Main Card Cage Main Card Cage (access) Card Cage (upper back plane board) (user I/O custom backplane) Card Cage Backplane (upper terminal blocks) Connector Function CNN1 AC/DC DRV CNN2 SMD CNN3 SMD CNN4 ISE CNN5 Optics CNN7 AC SENS CNN8 SENS 232 LVL CNN9 SM SENS CNN12 BCR SHUT TRG Card Cage (lower back plane board) Main Card Cage Block Diagram A Main Card Cage Block Diagram B Card Cage Backplane Cable Functions Schematic Diagram Card Cage Backplane Cable Diagram Link 7A_9009a.pdf Card Cage Backplane Cable Functions Backplane Cable Card Cage Slot & Board Destination Board and Connector Function CNN1 Slot 6 AC/DC Slot 5 SM AC/DC AC/DC Driver Bd CNN1 Pumps, Valves, AC Motors CNN2 Slot 4 SMC Slot 5 SM AC/DC SMD Bd CNN1 Motors (ISE, STR, Spl & Rgt Pumps, R1A Pip Arm) CNN3 Slot 4 SMC Slot 5 SM AC/DC SMD Bd CNN1 CNN4 Slot 1 DAQ ISE PRE Bd CNN5 Slot 1 DAQ Photo Pre Amp Bd CNN7 Slot 6 AC/DC CNN Bd CNN2 Sensors (Pumps Lids, Float Switches, STR) CNN8 Slot 5 SM AC/DC CNN Bd CNN4 “B”-Line Sensors, Valve FB Sensors, STR CUV Sensors CNN9 Slot 4 SMC CNN Bd CNN3 “A”- Line Sensors, Rgt 1 Crsl (Inner/Outer) Rgt 2 Carsl (Inner) CNN10 Slot 3 PM2 CNN Bd CNN25 Pressure Sensors Wash Cup Sensors CNN11 Slot 2 PM1 CNN Bd CNN22 Pressure Sensors Wash Cup Sensors CNN12 Slot 1 DAQ CNN Bd CNN21 BCR, LUI, CLI CNN14 Slot 5 SM AC/DC CNN Bd CNN20, 21, 22 (not used) Card Cage Backplane Cable Signals CNN1 SLOT SLOT6 (AC/DC) (To AC/DC Driver CN1) Pin No Pin NO. Signal Name Active I/O Note J5-A8 WSMUP 1 WS-UP L O J5-A7 WSMDOWN 2 WS-DOWN L O J5-A6 WSMBK 3 WS-BK L O J5-A5 STEMUP 4 STRZ-UP L O J5-A4 STEMDOWN 5 STRZ-DOWN L O J5-A3 STEMBK 6 STRZ-BK L O J5-A2 PWPMUP 7 PWP-UP L O J5-A1 PWPMDOWN 8 PWP-DOWN L O J4-A25 PWPMBK 9 PWP-BK L O J4-E1 6 AC1UP(GH) 10 PWBP-UP L O J4-E1 7 AC1DOWN(PS) 11 PWBP-DOWN L O J4-E1 8 AC1BK 12 PWBP-BK L O J5-B8 AUX1UP 13 AUXAC1-CW L O J5-B7 AUX1DOWN 14 AUXAC1-CCW L O J5-B6 AUX1BK 15 AUXAC1-BK L O SLOT6 (AC/DC) J5-B5 DPMUD 16 DTGP-ON L O J5-B4 DPMBK 17 DTGP-BK L O SLOT6 (AC/DC) J5-B3 CWPMUD 18 WCP-ON L O J5-B2 CWPMBK 19 WCP-BK L O SLOT6 (AC/DC) J5-C8 ISPMUD 20 ICTP-ON L O J5-C7 ISPMBK 21 ICTP-BK L O SLOT6 (AC/DC) J5-C6 IMPMUD 22 ICTR-ON L O J5-C5 IMPMBK 23 ICTR-BK L O SLOT6 (AC/DC) J5-C4 AUX2UD 24 AUXAC2-ON L O J5-C3 AUX2BK 25 AUXAC2-BK L O SLOT6 (AC/DC) J5-C2 CWVPMON 26 VP2 L O LCW Vacuum Pump AB J5-C1 CDVPMON 27 VP1 L O Cell Dryer Vacuum Pump AB J4-C25 FP1ON 28 TWMP L O DIW SUpply Pump1 (Water Bath) SLOT6 (AC/DC) SLOT6 (AC/DC) SLOT5 (SM_AC/DC) SLOT6 (AC/DC) SLOT6 (AC/DC) Cell Wash Stirrer Up/Down Probe Wash Pump (Inner Wall) A Probe Wash Pump (Inner Wall) B (C16000 only) AC Spare (CW/CCW) Detergent Pump Cell Wash Pump ISE Aspiration Pump ISE Reference Pump AC Spare (One way rotation) J4-C24 FP2ON 29 WPMP L O DIW Supply Pump2 (Outer Wall) J5-D8 MC1ON 30 MC1 L O Freezer Power Control Relay J5-D6 AUX3ON 31 AUXAC3-ON L O AC Spare (On/Off) J5-D5 ISPVON 32 IAVP L O ISE Aspiration Pump Mag Valve J5-D4 FV1ON 33 ESPV L O DIW Valve1 (S Outer Wall) J5-D3 FV2ON 34 ERPV L O DIW Valve2 (R Outer Wall) J5-D2 FV3ON 35 SWV L O DIW Valve3 (STR) J5-D1 FV4ON 36 WBSV L O DIW Valve4 (Water Bath Supply) J4-D25 FV5ON 37 PWV L O DIW Valve5 (Pressurized DIW Control) SLOT6(AC/DC) J5-E8 CWVON 38 CWV L O Cell Wash Water Valve SLOT6(AC/DC) J5-E7 DTVB14ON 39 DTG1V L O ALK&ACID Detergent Valve1 J5-E6 DTVB3ON 40 DTG2V L O ALK&ACID Detergent Valve2 SLOT6(AC/DC) J5-E5 DRYVON 41 CDV L O Cell Dryer Vac Pump Valve SLOT6(AC/DC) J5-E4 SPVON 42 ISPV L O Sampling Pump A Mag Valve SLOT6(AC/DC) J5-E3 RP1VON 43 IR1PV L O Reagent Pump 1A Mag Valve SLOT6(AC/DC) J5-E2 RP2VON 44 IR2PV L O Reagent Pump 2A Mag Valve SLOT5(SM_AC/DC) J5-E3 DC1 45 ISPBV L O Sampling Pump B Mag Valve (C16000 only) SLOT5(SM_AC/DC) J5-E4 DC2 46 IR1BPV L O Reagent Pump 1B Mag Valve (C16000 only) SLOT5(SM_AC/DC) J5-E5 DC3 47 IR2BPV L O Reagent Pump 2B Mag Valve (C16000 only) SLOT6(AC/DC) J4-E21 BUZZER 48 BUZZER L O Buzzer SLOT6(AC/DC) J4-E24 AUXMCW 49 IREF L O IREFCUP Sensor SLOT6(AC/DC) J4-E25 AUXVON 50 AUXDC1 L O Spare Mag Valve 51 GND - O 52 GND - O 55 +5V - O 56 +5V - O 57 +5V - O 58 +5V - O 59 +5V - O SLOT6 (AC/DC) SLOT6 (AC/DC) 53 54 60 CNN2 SLOT SLOT4 (SMC) SLOT4 (SMC) SLOT4 (SMC) SLOT4 (SMC) SLOT5 (SM_AC/DC) SLOT5 (SM_AC/DC) SLOT4 (SMC) SLOT5 (SM_AC/DC) +5V - O (To SMD CN1) Pin No Pin NO. Signal Name Active I/O J5-C3 IAAINA 1 IAHINA L O J5-C1 IAAINB 2 IAHINB L O J5-C2 IAATDA 3 IAHTDA L O J4-C25 IAATDB 4 IAHTDB L O J4-C24 IAAHOLD 5 IAHHOLD L O J5-C8 IAEINA 6 IAZINA L O J5-C6 IAEINB 7 IAZINB L O J5-C7 IAETDA 8 IAZTDA L O J5-C5 IAETDB 9 IAZTDB L O J5-C4 IAEHOLD 10 IAZHOLD L O J4-C23 STA1INA 11 STR1HINA L O J4-C21 STA1INB 12 STR1HINB L O J4-C22 STA1TDA 13 STR1HTDA L O J4-C20 STA1TDB 14 STR1HTDB L O J4-C19 STA1HOLD 15 STR1HHOLD L O J4-C18 STA2INA 16 STR2HINA L O J4-C16 STA2INB 17 STR2HINB L O J4-C17 STA2TDA 18 STR2HTDA L O J4-C15 STA2TDB 19 STR2HTDB L O J5-E17 STA2HOLD 20 STR2HHOLD L O J4-A17 SM1INA 21 STR1BHINA L O J4-A18 SM1INB 22 STR1BHINB L O J4-A19 SM1TDA 23 STR1BHTDA L O J4-A20 SM1TDB 24 STR1BHTDB L O J4-A21 SM1HOLD 25 STR1BHHOLD L O J4-A22 SM2INA 26 STR2BHINA L O J4-A23 SM2INB 27 STR2BHINB L O J4-A24 SM2TDA 28 STR2BHTDA L O J4-A25 SM2TDB 29 STR2BHTDB L O J5-A1 SM2HOLD 30 STR2BHHOLD L O J4-A20 SPINA 31 SPINA L O J4-A18 SPINB 32 SPINB L O J4-A19 SPTDA 33 SPTDA L O J4-A17 SPTDB 34 SPTDB L O J4-A16 SPHOLD 35 SPHOLD L O J5-A4 SM4INA 36 SPBINA L O J5-A5 SM4INB 37 SPBINB L O J5-A6 SM4TDA 38 SPBTDA L O J5-A7 SM4TDB 39 SPBTDB L O J5-A8 SM4HOLD 40 SPBHOLD L O Note ISE Arm Rotation Motor ISE Arm Up/Down Motor Stirrer 1A Horizontal Motor Stirrer 2A Horizontal Motor Stirrer 1B Horizontal Motor (C16000 only) Stirrer 2B Horizontal Motor (C16000 only) Sampling Pump A Up/Down Motor Sampling Pump B Up/Down Motor (C16000 only) SLOT4 (SMC) SLOT5 (SM_AC/DC) SLOT4 (SMC) SLOT5 (SM_AC/DC) SLOT4 (SMC) SLOT4 (SMC) J4-A15 R1PINA 41 R1PINA L O J5-C16 R1PINB 42 R1PINB L O J5-C17 R1PTDA 43 R1PTDA L O J5-C15 R1PTDB 44 R1PTDB L O J5-C14 R1PHOLD 45 R1PHOLD L O J5-A9 SM5INA 46 R1BPINA L O J5-A10 SM5INB 47 R1BPINB L O J5-A11 SM5TDA 48 R1BPTDA L O J5-A12 SM5TDB 49 R1BPTDB L O J5-A13 SM5HOLD 50 R1BPHOLD L O J5-C13 R2PINA 51 R2PINA L O J5-C11 R2PINB 52 R2PINB L O J5-C12 R2PTDA 53 R2PTDA L O J5-C10 R2PTDB 54 R2PTDB L O J5-C9 R2PHOLD 55 R2PHOLD L O J5-A14 SM6INA 56 R2BPINA L O J5-A15 SM6INB 57 R2BPINB L O J5-A16 SM6TDA 58 R2BPTDA L O J5-A17 SM6TDB 59 R2BPTDB L O J5-A18 SM6HOLD 60 R2BPHOLD L O J5-A10 R1AAINA 61 R1AHINA L O J5-A8 R1AAINB 62 R1AHINB L O J5-A9 R1AATDA 63 R1AHTDA L O J5-A7 R1AATDB 64 R1AHTDB L O J5-A6 R1AAHOLD 65 R1AHHOLD L O J5-A15 R1AEINA 66 R1AZINA L O J5-A13 R1AEINB 67 R1AZINB L O J5-A14 R1AETDA 68 R1AZTDA L O J5-A12 R1AETDB 69 R1AZTDB L O J5-A11 R1AEHOLD 70 R1AZHOLD L O 71 GND - O 72 GND - O 75 +5V - O 76 +5V - O 77 +5V - O 78 +5V - O 79 +5V - O 80 +5V - O R1 Pump A Up/Down Motor R1 Pump B Up/Down Motor (C16000 only) R2 Pump A Up/Down Motor R2 Pump B Up/Down Motor (C16000 only) R1 Arm A Rotation Motor R1 Arm A Up/Down Motor 73 74 CNN3 I/O Connector SLOT (To SMD CN1) Pin No Pin NO. Signal Name Active I/O Note SLOT4 (SMC) SLOT4 (SMC) SLOT5 (SM_AC/DC) SLOT5 (SM_AC/DC) SLOT5 (SM_AC/DC) SLOT5 (SM_AC/DC) SLOT4 (SMC) SLOT4 (SMC) SLOT5 (SM_AC/DC) J4-A25 R2AAINA 1 R2AHINA L O J4-A23 R2AAINB 2 R2AHINB L O J4-A24 R2AATDA 3 R2AHTDA L O J4-A22 R2AATDB 4 R2AHTDB L O J4-A21 R2AAHOLD 5 R2AHHOLD L O J5-A5 R2AEINA 6 R2AZINA L O J5-A3 R2AEINB 7 R2AZINB L O J5-A4 R2AETDA 8 R2AZTDA L O J5-A2 R2AETDB 9 R2AZTDB L O J5-A1 R2AEHOLD 10 R2AZHOLD L O J4-B18 SM7INA 11 R1ABHINA L O J4-B19 SM7INB 12 R1ABHINB L O J4-B20 SM7TDA 13 R1ABHTDA L O J4-B21 SM7TDB 14 R1ABHTDB L O J4-B22 SM7HOLD 15 R1ABHHOLD L O J5-B4 SM8INA 16 R1ABZINA L O J5-B5 SM8INB 17 R1ABZINB L O J5-B6 SM8TDA 18 R1ABZTDA L O J5-B7 SM8TDB 19 R1ABZTDB L O J5-B8 SM8HOLD 20 R1ABZHOLD L O J5-B9 SM9INA 21 R2ABHINA L O J5-B10 SM9INB 22 R2ABHINB L O J5-B11 SM9TDA 23 R2ABHTDA L O J5-B12 SM9TDB 24 R2ABHTDB L O J5-B13 SM9HOLD 25 R2ABHHOLD L O J5-B14 SM10INA 26 R2ABZINA L O J5-B15 SM10INB 27 R2ABZINB L O J5-B16 SM10TDA 28 R2ABZTDA L O J5-B17 SM10TDB 29 R2ABZTDB L O J5-B18 SM10HOLD 30 R2ABZHOLD L O J5-E16 ISIINA 31 LSHINA L O J5-E14 ISIINB 32 LSHINB L O J5-E15 ISITDA 33 LSHTDA L O J5-E13 ISITDB 34 LSHTDB L O J5-E12 ISIHOLD 35 LSHHOLD L O J5-E11 ISOINA 36 PPHCINA L O J5-E9 ISOINB 37 PPHCINB L O J5-E10 ISOTDA 38 PPHCTDA L O J5-E8 ISOTDB 39 PPHCTDB L O J5-E7 ISOHOLD 40 PPHCHOLD L O J4-C21 SM12INA 41 PPBHCINA L O J4-C22 SM12INB 42 PPBHCINB L O J4-C23 SM12TDA 43 PPBHCTDA L O J4-C24 SM12TDB 44 PPBHCTDB L O R2 Arm A Rotation Motor R2 Arm A Up/Down Motor R1 Arm B Rotation Motor (C16000 only) R1 Arm B Up/Down Motor (C16000 only) R2 Arm B Rotation Motor (C16000 only) R2 Arm B Up/Down Motor (C16000 only) LSH Rotation Motor (LSH Inner Disk Rotation) (C8000 & C16000 only) HCW Pump A Motor (LSH Outer Disk Rotation) HCW Pump B Motor (Rack Sampler Lead Motor) (C16000 only) SLOT4 (SMC) SLOT4 (SMC) SLOT4 (SMC) SLOT4 (SMC) SLOT4 (SMC) SLOT5 (SM_AC/DC) SLOT4 (SMC) SLOT4 (SMC) SLOT5 (SM_AC/DC) SLOT4 (SMC) J4-C25 SM12HOLD 45 PPBHCHOLD L O J5-E6 AUXINA 46 AUXINA L O J5-E4 AUXINB 47 AUXINB L O J5-E5 AUXTDA 48 AUXTDA L O J5-E3 AUXTDB 49 AUXTDB L O J5-E2 AUXHOLD 50 AUXHOLD L O J5-E1 RUCW 51 RCCW L O J4-E25 RUCCW 52 RCCCW L O J4-E24 R1CICW 53 R1ICCW L O J4-E23 R1CICCW 54 R1ICCCW L O J4-E22 R1COCW 55 R1OCCW L O J4-E21 R1COCCW 56 R1OCCCW L O J4-E20 R2CCW 57 R2ICCW L O J4-E19 R2CCCW 58 R2ICCCW L O J5-A2 SM3CW 59 R2OCCW L O J5-A3 SM3CCW 60 R2OCCCW L O J4-E18 SAACW 61 SAHCW L O J4-E17 SAACCW 62 SAHCCW L O J4-E16 SAECW 63 SAZCW L O J4-E15 SAECCW 64 SAZCCW L O J4-B24 SM13CW 65 SPCW L O J4-B25 SM13CCW 66 SPCCW L O J5-A17 ACW 67 AUXCW L O J5-A16 ACCW 68 AUXCCW L O 71 GND - O 72 GND - O 75 +5V - O 76 +5V - O 77 +5V - O 78 +5V - O 79 +5V - O 80 +5V - O 86 SM11INA L O 87 SM11INB L O Spare Reaction Cuvette Rotation Motor R1 Inner Disk Rotation Motor R1 Outer Disk Rotation Motor R2 Inner Disk Rotation Motor R2 Outer Disk Rotation Motor (C16000 only) Sampling Arm Rotation Motor Sampling Arm Up/Down Motor (Sampling Pump A Up/Down Motor) Spare 69 70 73 74 81 82 83 84 85 SLOT5 (SM_AC/DC) J4-C16 SM11INA J4-C17 SM11INB 200FR Spare Motor J4-C18 SM11TDA 88 SM11TDA L O J4-C19 SM11TDB 89 SM11TDB L O J4-C20 SM11HOLD 90 SM11HOLD L O J5-C4 SM13INA 91 SM13INA L O J5-C5 SM13INB 92 SM13INB L O J5-C6 SM13TDA 93 SM13TDA L O J5-C7 SM13TDB 94 SM13TDB L O J5-C8 SM13HOLD 95 SM13HOLD L O Spare Motor 96 97 98 99 100 CNN4 I/O Connector (To ISE PRE) SLOT SLOT1 DAQ Pin No. No Active I/O J5-A14 NaOUT 1 Na OUT AI J5-B14 NaOUTG 2 Na GND AI J5-A13 KOUT 3 K OUT AI J5-B13 KOUTG 4 K GND AI J5-E13 CLOUT 5 CL OUT AI J5-D13 CLOUTG 6 CL GND AI 9 +15V O 10 AGND O 7 8 11 12 13 14 CNN5 I/O Connector (To Photo Pre-Amp) SLOT SLOT1 DAQ Pin No No Active I/O - AGND 1 AGND(±15VGND) O - AGND 2 AGND(±15VGND) O J5-A22 340nm 3 340nm AI J5-B22 340nmG 4 340nmG AI J5-A21 380nm 5 380nm AI J5-B21 380nmG 6 380nmG AI J5-A20 404nm 7 404nm AI J5-B20 404nmG 8 404nmG AI J5-A19 412nm 9 412nm AI J5-B19 412nmG 10 412nmG AI J5-A18 444nm 11 444nm AI J5-B18 444nmG 12 444nmG AI J5-A17 476nm 13 476nm AI J5-B17 476nmG 14 476nmG AI J5-A16 500nm 15 500nm AI J5-B16 500nmG 16 500nmG AI J5-A15 524nm 17 524nm AI J5-B15 524nmG 18 524nmG AI J5-E22 548nm 19 548nm AI J5-D22 548nmG 20 548nmG AI J5-E21 572nm 21 572nm AI J5-D21 572nmG 22 572nmG AI J5-E20 604nm 23 604nm AI J5-D20 604nmG 24 604nmG AI J5-E19 628nm 25 628nm AI J5-D19 628nmG 26 628nmG AI J5-E18 660nm 27 660nm AI J5-D18 660nmG 28 660nmG AI J5-E17 700nm 29 700nm AI J5-D17 700nmG 30 700nmG AI J5-E16 748nm 31 748nm AI J5-D16 748nmG 32 748nmG AI J5-E15 804nm 33 804nm AI J5-D15 804nmG 34 804nmG AI 35 +15V O 36 +15V O 37 +15V O 38 -15V O 39 -15V O 40 -15V O CNN7 I/O Connector SLOT SLOT6 (AC/DC) (CNN Board CNN2) Pin No Pin NO. Signal Name Active I/O J4-A1 WSUP 1 WSUP H I J4-B1 WSDOWN 2 WSDOWN H I J4-A2 WSDOWNOK 3 WSDOWNOK H I J4-B2 STEUP 4 STEUP H I J4-A3 STEDOWN 5 STEDOWN H I J4-B3 STEDOK 6 STEDOK H I J4-A6 PWPUP 7 PWPUP H I J4-B6 PWPDOWN 8 PWPDOWN H I J4-B9 AC1UP 9 AC1UP H I J4-A10 AC1DOWN 10 AC1DOWN H I J4-B10 AC1OK 11 AC1OK H I Note Cell Wash Up/Down Stirrer Up/Down Probe Wash Pump (Inner Wall) AC Spare (CW/CCW) J4-A4 DPUP 12 DPUP H I Detergent Pump J4-B4 DPDOWN 13 DPDOWN H I J4-A5 CWPUP 14 CWPUP H I J4-B5 CWPDOWN 15 CWPDOWN H I J4-A8 ISPUP 16 ISPUP H I J4-B8 ISPDOWN 17 ISPDOWN H I J4-A9 IMPUP 18 IMPUP H I ISE Reference Pump J4-A11 AC2UP 19 AC2UP H I J4-B11 AC2DOWN 20 AC2DOWN H I AC Spare (One Way Direction) J4-A7 TWPUP 21 TWPUP H I J4-B7 TWPDOWN 22 TWPDOWN H I J4-E1 ST1ON 23 ST1ON L O J4-E2 ST1SCAN 24 ST1SCAN L O J4-E3 ST1CLR 25 ST1CLR L O J4-E7 ST1BUSY0 26 ST1BUSY0 L I J4-E8 ST1BUSY1 27 ST1BUSY1 L I J4-E4 ST2ON 28 ST2ON L O J4-E5 ST2SCAN 29 ST2SCAN L O J4-E6 ST2CLR 30 ST2CLR L O J4-E9 ST2BUSY0 31 ST2BUSY0 L I J4-E10 ST2BUSY1 32 ST2BUSY1 L I J4-E11 +5V (IN) 33 +5V (IN) H I J4-C1 ISLED1 34 ISLED1 L O J4-D1 ISLED2 35 ISLED2 L O J4-C2 ISLUILED1 36 ISLUILED1 L O J4-D2 ISLUILED2 37 ISLUILED2 L O J4-C3 AUXLED1 38 AUXLED1 L O J4-D3 AUXLED2 39 AUXLED2 L O J4-C4 CPULED0 40 CPULED0 L O J4-D4 CPULED1 41 CPULED1 L O J4-C5 CPULED2 42 CPULED2 L O J4-D5 CPULED3 43 CPULED3 L O J4-C6 CPULED4 44 CPULED4 L O J4-D6 CPULED5 45 CPULED5 L O J4-C7 CPULED6 46 CPULED6 L O J4-D7 CPULED7 47 CPULED7 L O J4-C8 CPULED8 48 CPULED8 L O J4-D8 CPULED9 49 CPULED9 L O J4-C9 CPULED10 50 CPULED10 L O J4-D9 CPULED11 51 CPULED11 L O J4-C10 CPULED12 52 CPULED12 L O J4-D10 CPULED13 53 CPULED13 L O J4-C11 CPULED14 54 CPULED14 L O Cell Wash Pump ISE Aspiration Pump HCW Pump A STR DRV ISLED IS-LUI-LED AUX-LED Status LED J4-D11 CPULED15 55 CPULED15 L O J4-A15 AUXMUP 56 AUXMUP H I J4-C15 AUXMDOWN 57 AUXMDOWN H I J4-B15 BATHLVL 58 BATHLVL H I Water Bath Level Sensor J4-B16 R1CCVROK 59 R1CCVROK H I R1 Cover OK Sensor J4-A16 R2CCVROK 60 R2CCVROK H I R2 Cover OK Sensor J4-B18 ISCVROK 61 ISCVROK H I ISH Cover OK Sensor J4-C16 FFS1 62 FFS1 I FFS1: FSDT Drain Tank (LCW) J4-A17 FFS2 63 FFS2 I FFS2: FSLU Water Tank Upper Limit J4-B17 FFS3 64 FFS3 I FFS3: FSLH Water Tank Level Hi J4-C17 FFS4 65 FFS4 I FFS4: FSLL Water Tank Level Low J4-A18 FFS5 66 FFS5 I FFS5: Spare J4-C18 DTF 67 DTF I DTF: Drain Tank (HCW) J4-A22 DALARM 68 DALARM I Degasser J4-A23 DALARMV 69 DALARMV I J4-C19 ESMCC 70 ESMCC H I Reference Solution J4-A19 ESB14 71 ESB14 H I Alkaline Detergent J4-B20 ESB3 72 ESB3 H I Acid Detergent J4-B22 ESMCCSS 73 ESMCCSS_IREFCUP H I IREFCUP Sensor J4-B23 ESB14SS 74 ESB14SS H I Alkaline Detergent Sensor J4-B24 ESB3SS 75 ESB3SS H I Acid Detergent Sensor J4-D15 IPMON 76 IPMON L O Water Bath Cycle Pump J4-D16 IHTRON 77 IHTRON L O Water Bath Heater J4-E16 IMVOPEN 78 IMVOPEN L O J4-E15 IMVSHUT 79 IMVSHUT L O Water Bath Motor Valve J4-E17 IFANON 80 IFANON L O Water Bath Fan Control J5-B14 ISTEMP+ 81 ISTEMP+ AI J5-B15 ISTEMP- 82 ISTEMP- AI LSH Temp Varistor J5-E14 BATHTMP+ 83 BATHTMP+ AI J5-E15 BATHTMP- 84 BATHTMP- AI J5-D14 R1CTEMP+ 85 R1CTEMP+ AI J5-D15 R1CTEMP- 86 R1CTEMP- AI J5-C14 R2CTEMP+ 87 R2CTEMP+ AI J5-C15 R2CTEMP- 88 R2CTEMP- AI J5-A14 IVRSTR+ 89 IVRSTR+ AI Spare DC Motor Sensor Cuvette Temp Varistor R1 Supply Temp Varistor R2 Supply Temp Varistor Water Bath Fan IVRSTR- 90 IVRSTR- AI J4-B19 ISSW 91 ISSW I IA-LUI SW J4-C22 ISLSW 92 ISLSW I IALSW SW J4-B21 AUXSW1 93 AUXSW1 I Spare SW1 J4-C21 AUXSW2 94 AUXSW2 I Spare SW2 J4-A20 CUPSNSRI 95 CUPSNSRI I Spare Sensor J4-C20 CUPSNSRO 96 CUPSNSRO I Spare Sensor GND 97 GND O GND GND 98 GND O GND GND 99 GND O GND GND 100 GND O GND CNN8 SLOT SLOT5 (SM_AC/DC) Control Varistor J5-A15 (To CNN Board CNN4) Pin No Pin NO. Signal Name Active I/O Note J4-A1 SM1HOME 1 SM1HOME H I 1B Stirrer Horizontal Motor (C16000 only) J4-A2 SM2HOME 2 SM2HOME H I 2B Stirrer Horizontal Motor (C16000 only) J4-A5 SM4HOME 3 SM4HOME H I J4-A6 SM4PITCH 4 SM4PITCH H I Sampling Pump B Up/Down Motor (C16000 only) J4-A7 SM4DLMT 5 SM4DLMT H I J4-A8 SM5HOME 6 SM5HOME H I J4-A9 SM5PITCH 7 SM5PITCH H I J4-A10 SM5DLMT 8 SM5DLMT H I J4-A11 SM6HOME 9 SM6HOME H I J4-A15 SM6PITCH 10 SM6PITCH H I J4-A16 SM6DLMT 11 SM6DLMT H I J4-B1 SM7HOME 12 SM7HOME H I J4-B2 SM7PITCH 13 SM7PITCH H I J4-B3 SM8HOME 14 SM8HOME H I J4-B4 SM8PITCH 15 SM8PITCH H I J4-B5 SM8CLASH 16 SM8CLASH H I J4-B6 SM9HOME 17 SM9HOME H I J4-B7 SM9PITCH 18 SM9PITCH H I J4-B8 SM10HOME 19 SM10HOME H I J4-B9 SM10PITCH 20 SM10PITCH H I J4-B10 SM10CLASH 21 SM10CLASH H I J4-C9 SM12HOME 22 SM12HOME H I J4-C10 SM12PITCH(OK) 23 SM12PITCH(OK) H I J4-A3 SM3HOME 24 SM3HOME H I J4-A4 SM3PITCH 25 SM3PITCH H I J4-C11 SM13HOME 26 SM13HOME H I R1 Pump B Up/Down Motor (C16000 only) R2 Pump B Up/Down Motor (C16000 only) R1 Arm B Rotate Motor (C16000 only) R1 Arm B Up/Down Motor (C16000 only) R2 Arm B Rotate Motor (C16000 only) R2 Arm B Up/Down Motor (C16000 only) HCW Pump B Motor (C16000 only) R2 Outer Disk (C16000 only) Rotate Motor Sampling Pump A Up/Down Motor J4-C15 SM13PITCH(OK) 27 SM13PITCH(OK) H I J4-B23 SM13DLMT 28 SM13DLMT H I J4-B11 SM11HOME 29 SM11HOME H I J4-B15 SM11PITCH 30 SM11PITCH H I J4-B16 SM11B 31 SM11B H I J4-B17 SM11F 32 SM11F H I J4-E1 AC1UP(HOME) 33 AC1UP(HOME) H I J4-E2 AC1DOWN(MOVE) 34 AC1DOWN(MOVE) H I J4-E3 AC1S9 35 AC1S9 H I J4-E4 AC1OK 36 AC1OK H I J4-E10 SENS1 37 SENS1 H I STR1A CUV Sensor J4-E11 SENS2 38 SENS2 H I STR2A CUV Sensor J4-E15 SENS3 39 SENS3 H I STR1B CUV Sensor J4-D1 SENS4 40 SENS4 H I STR2B CUV Sensor J4-D2 SENS5 41 SENS5 H I ISE Aspiration Pump Mag Valve FB Sig. J4-D3 SENS6 42 SENS6 H I DIW Valve2 (R Outer) FB Sig. J4-D4 SENS7 43 SENS7 H I DIW Valve1 (S Outer) FB Sig. J4-D5 SENS8 44 SENS8 H I DIW Valve2 (R Outer) FB Sig. J4-D6 SENS9 45 SENS9 H I DIW Valve4 (Water Bath Supply) FB Sig. J4-D7 SENS10 46 SENS10 H I DIW Valve 5 (Pressurized Water Control) FB Sig. J4-D8 SENS11 47 SENS11 H I Cell Wash Water Mag Valve FB Sig. J4-D9 SENS12 48 SENS12 H I ALK&ACID Detergent Mag Valve1 FB Sig. J4-D10 SENS13 49 SENS13 H I ALK&ACID Detergent Mag Valve2 FB Sig. J4-D11 SENS14 50 SENS14 H I Cell Dryer Vacuum Pump Mag Valve FB Sig. J4-D15 SENS15 51 SENS15 H I Sampling Pump A Mag Valve FB Sig. J4-C1 SENS16 52 SENS16 H I Reagent Pump 1A Mag Valve FB Sig. J4-C2 SENS17 53 SENS17 H I Reagent Pump 2A Mag Valve FB Sig. J4-C3 SENS18 54 SENS18 H I Sampling Pump B Mag Valve FB Sig. J4-C4 SENS19 55 SENS19 H I Reagent Pump 1B Spare Motor Probe Wash Pump (Inner Wall) B (C16000 only) Mag Valve FB Sig. J4-C5 SENS20 56 SENS20 H I Reagent Pump 2B Mag Valve FB Sig. J4-C6 SENS21 57 SENS21 H I Buzzer FB Sig. J4-C7 SENS22 58 SENS22 H I IREFCUP Sensor FB Sig. J4-C8 SENS23 59 SENS23 H I Spare Mag Valve FB Sig. J5-D15 FB12VGND 60 FB12VGND L I FB12V GND J5-C9 SENS24 61 SENS24 L I LCW Vacuum Pump AB FB Sig. J5-C10 SENS25 62 SENS25 L I LCW Vacuum Pump AB FB Sig. J5-C11 SENS26 63 SENS26 L I DIW Supply Pump 1 (Water Bath) FB Sig. J5-C12 SENS27 64 SENS27 L I DIW Supply Pump 2 (Outer Wall Wash Pump) FB Sig. J5-C13 SENS28 65 SENS28 L I Freezer Power Control Relay FB Sig. J5-C14 SENS29 66 SENS29 L I AC Spare (ON/OFF) FB Sig. J5-A19 VCC(LVL1) 67 VCC(LVL1) I J5-A20 LVL1 68 LVL1 R1B Level Sensor (C16000 only) J5-A21 VCC(LVL1OK) 69 VCC(LVL1OK) J5-A22 LVL1OK 70 LVL1OK J5-C21 VCC(AUTO0_1) 71 VCC(AUTO0_1) J5-C22 AUTO0_1 72 AUTO0_1 J5-B19 VCC(LVL2) 73 VCC(LVL2) J5-B20 LVL2 74 LVL2 J5-B21 VCC(LVL2OK) 75 VCC(LVL2OK) J5-B22 LVL2OK 76 LVL2OK J5-C19 VCC(AUTO0_2) 77 VCC(AUTO0_2) J5-C20 AUTO0_2 78 AUTO0_2 L J5-D16 DC30(TRIG1) 79 DC30(TRIG1) L O BCR1 (RS232C-1) J5-D17 DTR1 80 DTR1 J5-D18 DSR1 81 DSR1 J5-D19 CTS1 82 CTS1 J5-D20 RTS1 83 RTS1 J5-D21 RXD1 84 RXD1 J5-D22 TXD1 85 TXD1 J5-E16 DC31(TRIG2) 86 DC31(TRIG2) L O BCR2 (RS232C-1) J5-E17 DTR2 87 DTR2 J5-E18 DSR2 88 DSR2 J5-E19 CTS2 89 CTS2 J5-E20 RTS2 90 RTS2 H I I H I L I H I R2B Level Sensor (C16000 only) I H I J5-E21 RXD2 91 RXD2 J5-E22 TXD2 92 TXD2 93 94 95 96 GND O 97 GND O 98 GND O 99 GND O 100 GND O CNN9 SLOT SLOT4 (SMC) (To CNN Board CNN3) Pin No Pin NO. Signal Name Active I/O Note J4-C1 IAAHOME 1 IAAHOME H I ISE Arm Rotate Motor J4-D1 IAEHOME 2 IAEHOME H I J4-A2 IAEM1R 3 IAEM1R H I ISE Arm Up/Down Motor J4-B2 STA1HOME 4 STA1HOME H I Stirrer 1A Horizontal Motor J4-C2 STA2HOME 5 STA2HOME H I Stirrer 2A Horizontal Motor J4-C5 SPHOME 6 SPHOME H I J4-D5 SPM1R 7 SPM1R H I Sampling Pump A Up/Down Motor J4-A6 SPDLMT 8 SPDLMT H I J4-B6 R1PHOME 9 R1PHOME H I J4-C6 R1PM1R 10 R1PM1R H I J4-D6 R1PDLMT 11 R1PDLMT H I J4-A7 R2PHOME 12 R2PHOME H I J4-B7 R2PM1R 13 R2PM1R H I J4-C7 R2PDLMT 14 R2PDLMT H I J4-D7 R1AAHOME 15 R1AAHOME H I J4-A8 R1AAM1R 16 R1AAM1R H I J4-B8 R1AEHOME 17 R1AEHOME H I J4-C8 R1AEM1R 18 R1AEM1R H I J4-D8 R1ACLASH 19 R1ACLASH H I J4-A9 R2AAHOME 20 R2AAHOME H I J4-B9 R2AAM1R 21 R2AAM1R H I J4-C9 R2AEHOME 22 R2AEHOME H I J4-D9 R2AEM1R 23 R2AEM1R H I J4-A10 R2ACLASH 24 R2ACLASH H I J4-B10 ISIHOME 25 ISIHOME H I J4-C10 ISIPITCH 26 ISIPITCH H I J4-E4 ISIPOS 27 ISIPOS H I J4-D10 ISOHOME 28 ISOHOME H I J4-A11 ISOPITCH 29 ISOPITCH H I R1 Pump A Up/Down Motor R2 Pump A Up/Down Motor R1 Arm A Rotate Motor R1 Arm A Up/Down Motor R2 Arm A Rotate Motor R2 Arm A Up/Down Motor LSH Rotate Disk Motor HCW Pump A Motor J4-E5 ISOPOS 30 ISOPOS H I J4-B11 AHOME 31 AHOME H I J4-C11 AM1R 32 AM1R H I J4-A1 RUHOME 33 RUHOME H I J4-B1 RUM1R 34 RUM1R H I J4-D2 R1CIHOME 35 R1CIHOME H I J4-A3 R1CIPITCH 36 R1CIPITCH H I J4-E1 R1CIPOS 37 R1CIPOS H I J4-B3 R1COHOME 38 R1COHOME H I J4-C3 R1COPITCH 39 R1COPITCH H I J4-E2 R1COPOS 40 R1COPOS H I J4-D3 R2CHOME 41 R2CHOME H I J4-A4 R2CPITCH 42 R2CPITCH H I J4-E3 R2CPOS 43 R2CPOS H I J4-B4 SAAHOME 44 SAAHOME H I J4-C4 SAAM1R 45 SAAM1R H I J4-D4 SAEHOME 46 SAEHOME H I J4-A5 SAEM1R 47 SAEM1R H I J4-B5 SACLASH 48 SACLASH H I J5-A19 VCC(SALVL) 49 VCC(SALVL) J5-A20 SALVL 50 SALVL J5-A21 VCC(SALVLOK) 51 VCC(SALVLOK) J5-A22 SALVLOK 52 SALVLOK J5-B19 VCC (SAAUTOZERO) 53 VCC (SAAUTOZERO) J5-B20 SAAUTOZERO 54 SAAUTOZERO J5-B21 VCC(R1ALVL) 55 VCC(R1ALVL) J5-B22 R1ALVL 56 R1ALVL J5-C19 VCC(R1ALVLOK) 57 VCC (R1ALVLOK) J5-C20 R1ALVLOK 58 R1ALVLOK J5-C21 VCC (R1AAUTOZERO) 59 VCC (R1AAUTOZERO) J5-C22 R1AAUTOZERO 60 R1AAUTOZERO J5-D19 VCC(R2ALVL) 61 VCC(R2ALVL) J5-D20 R2ALVL 62 R2ALVL J5-D21 VCC(R2ALVLOK) 63 VCC(R2ALVLOK) J5-D22 R2ALVLOK 64 R2ALVLOK J5-E19 VCC (R2AAUTOZERO) 65 VCC (R2AAUTOZERO) J5-E20 R2AAUTOZERO 66 R2AAUTOZERO J4-E7 SALVLGU 67 SALVLGU J4-E6 SALVLGUV(+5V) 68 SALVLGUV(+5V) 69 70 71 I H I Spare Cuvette Rotate Disk Motor R1 Inner Rotate Disk Motor R1 Outer Rotate Disk Motor R2 Inner Rotate Disk Motor Sampling Arm Rotate Motor Sampling Arm Up/Down Motor Sample Level Sensor I H I O L O I H R1 Level Sensor I I H I O L O I H I I H I O L L O R2 Level Sensor 72 73 74 75 76 GND O GND 77 GND O GND 78 GND O GND 79 GND O GND 80 GND O GND CNN10 SLOT SLOT3 (PM2) (To CNN Board CNN25) Pin No Pin NO. Signal Name Active I/O Note J5-C18 SASEN1_POS 1 SASEN1_POS I Sample Probe Sensor output (+) J5-C17 SASEN1_NEG 2 SASEN1_NEG I Sample Probe Sensor output (-) J5-A2 R1ASEN_POS 3 R1ASEN_POS I Reagent 1a Pressure Sensor output (+) J5-A1 R1ASEN_NEG 4 R1ASEN_NEG I Reagent 1a Pressure Sensor output (-) J5-A4 R2ASEN_POS 5 R2ASEN_POS I Reagent 2a Pressure Sensor output (+) J5-A3 R2ASEN_NEG 6 R2ASEN_NEG I Reagent 2a Pressure Sensor output (-) J5-D18 SASEN2_POS 7 SASEN2_POS I 2nd Sample Probe Sensor output (+) J5-D17 SASEN2_NEG 8 SASEN2_NEG I 2nd Sample Probe Sensor output (-) J5-A6 R1BSEN_POS 9 R1BSEN_POS I Reagent 1b Pressure Sensor output (+) J5-A5 R1BSEN_NEG 10 R1BSEN_NEG I Reagent 1b Pressure Sensor output (-) J5-A8 R2BSEN_POS 11 R2BSEN_POS I Reagent 2b Pressure Sensor output (+) J5-A7 R2BSEN_NEG 12 R2BSEN_NEG I Reagent 2b Pressure Sensor output (-) J5-C2 R1AWCSEN_POS 13 R1AWCSEN_POS I Reagent 1a Wash cup sensor output (+) J5-C1 R1AWCSEN_NEG 14 R1AWCSEN_NEG I Reagent 1a Wash cup sensor output (-) J5-C4 R2AWCSEN_POS 15 R2AWCSEN_POS I Reagent 2a Wash cup sensor output (+) J5-C3 R2AWCSEN_NEG 16 R2AWCSEN_NEG I Reagent 2a Wash cup sensor output (-) J5-C6 R1BWCSEN_POS 17 R1BWCSEN_POS I Reagent 1b Wash cup sensor output (+) J5-C5 R1BWCSEN_NEG 18 R1BWCSEN_NEG I Reagent 1b Wash cup sensor output (-) J5-C8 R2BWCSEN_POS 19 R2BWCSEN_POS I Reagent 2b Wash cup sensor output (+) J5-C7 R2BWCSEN_NEG 20 R2BWCSEN_NEG I Reagent 2b Wash cup sensor output (-) J5-C10 SAWCSEN1_POS 21 SAWCSEN1_POS I Sample Probe Wash cup sensor 1 output (+) J5-C9 SAWCSEN1_NEG 22 SAWCSEN1_NEG I Sample Probe Wash cup sensor 1 output (-) J5-C12 SAWCSEN2_POS 23 SAWCSEN2_POS I Sample Probe Wash cup sensor 2 output (+) J5-C11 SAWCSEN2_NEG 24 SAWCSEN2_NEG I Sample Probe Wash cup sensor 2 output (-) 25 AUXSENS1_POS I Aux1 Wash cup sensor output (+) 26 AUXSENS1_NEG I Aux1 Wash cup sensor output (-) 27 AUXSENS2_POS I Aux2 Wash cup sensor output (+) 28 AUXSENS2_NEG I Aux2 Wash cup sensor output (-) 29 AUXSENS3_POS I Aux2 Wash cup sensor output (+) 30 AUXSENS3_NEG I Aux2 Wash cup sensor output (-) 31 AUXSENS4_POS I Aux2 Wash cup sensor output (+) 32 AUXSENS4_NEG I Aux2 Wash cup sensor output (-) Signal Definition Only / Non Connection J5-E22 SLLS 33 SLLS I Sample LLS signal A4 on the c8000 LLS PCB J5-E21 SLLSG 34 SLLSG I Sample LLS signal ground, AG on the c8000 LLS PCB J5-E16 SACLASH 35 SACLASH I Sample Probe Crash Sensor output signal 36 +12V O 37 +12V O 38 +12V O 39 +12V O 40 +12V O 41 -12V O 42 -12V O 43 -12V O 44 -12V O 45 -12V O 46 ±12VGND O 47 ±12VGND O 48 ±12VGND O 49 ±12VGND O 50 ±12VGND O CNN11 SLOT SLOT2 (PM1) (To CNN Board CNN22) Pin No Pin NO. Signal Name Active I/O Note J5-C18 SASEN1_POS 1 SASEN1_POS I Sample Probe Sensor output (+) J5-C17 SASEN1_NEG 2 SASEN1_NEG I Sample Probe Sensor output (-) J5-A2 R1ASEN_POS 3 R1ASEN_POS I Reagent 1a Pressure Sensor output (+) J5-A1 R1ASEN_NEG 4 R1ASEN_NEG I Reagent 1a Pressure Sensor output (-) J5-A4 R2ASEN_POS 5 R2ASEN_POS I Reagent 2a Pressure Sensor output (+) J5-A3 R2ASEN_NEG 6 R2ASEN_NEG I Reagent 2a Pressure Sensor output (-) J5-D18 SASEN2_POS 7 SASEN2_POS I 2nd Sample Probe Sensor output (+) J5-D17 SASEN2_NEG 8 SASEN2_NEG I 2nd Sample Probe Sensor output (-) J5-A6 R1BSEN_POS 9 R1BSEN_POS I Reagent 1b Pressure Sensor output (+) J5-A5 R1BSEN_NEG 10 R1BSEN_NEG I Reagent 1b Pressure Sensor output (-) J5-A8 R2BSEN_POS 11 R2BSEN_POS I Reagent 2b Pressure Sensor output (+) J5-A7 R2BSEN_NEG 12 R2BSEN_NEG I Reagent 2b Pressure Sensor output (-) J5-C2 R1AWCSEN_POS 13 R1AWCSEN_POS I Reagent 1a Wash cup sensor output (+) J5-C1 R1AWCSEN_NEG 14 R1AWCSEN_NEG I Reagent 1a Wash cup sensor output (-) J5-C4 R2AWCSEN_POS 15 R2AWCSEN_POS I Reagent 2a Wash cup sensor output (+) J5-C3 R2AWCSEN_NEG 16 R2AWCSEN_NEG I Reagent 2a Wash cup sensor output (-) J5-C6 R1BWCSEN_POS 17 R1BWCSEN_POS I Reagent 1b Wash cup sensor output (+) J5-C5 R1BWCSEN_NEG 18 R1BWCSEN_NEG I Reagent 1b Wash cup sensor output (-) J5-C8 R2BWCSEN_POS 19 R2BWCSEN_POS I Reagent 2b Wash cup sensor output (+) J5-C7 R2BWCSEN_NEG 20 R2BWCSEN_NEG I Reagent 2b Wash cup sensor output (-) J5-C10 SAWCSEN1_POS 21 SAWCSEN1_POS I Sample Probe Wash cup sensor 1 output (+) J5-C9 SAWCSEN1_NEG 22 SAWCSEN1_NEG I Sample Probe Wash cup sensor 1 output (-) J5-C12 SAWCSEN2_POS 23 SAWCSEN2_POS I Sample Probe Wash cup sensor 2 output (+) J5-C11 SAWCSEN2_NEG 24 SAWCSEN2_NEG I Sample Probe Wash cup sensor 2 output (-) 25 AUXSENS1_POS I Aux1 Wash cup sensor output (+) 26 AUXSENS1_NEG I Aux1 Wash cup sensor output (-) 27 AUXSENS2_POS I Aux2 Wash cup sensor output (+) 28 AUXSENS2_NEG I Aux2 Wash cup sensor output (-) 29 AUXSENS3_POS I Aux2 Wash cup sensor output (+) 30 AUXSENS3_NEG I Aux2 Wash cup sensor output (-) 31 AUXSENS4_POS I Aux2 Wash cup sensor output (+) 32 AUXSENS4_NEG I Aux2 Wash cup sensor output (-) 33 SLLS I Sample LLS signal A4 on the c8000 LLS PCB Signal Definition Only / Non Connection J5-E22 SLLS J5-E21 SLLSG 34 SLLSG I Sample LLS signal ground, AG on the c8000 LLS PCB J5-E16 SACLASH 35 SACLASH I Sample Probe Crash Sensor output signal 36 +12V O 37 +12V O 38 +12V O 39 +12V O 40 +12V O 41 -12V O 42 -12V O 43 -12V O 44 -12V O 45 -12V O 46 ±12VGND O 47 ±12VGND O 48 ±12VGND O 49 ±12VGND O 50 ±12VGND O CNN12 SLOT SLOT1 (DAQ) (To CNN Board CNN20, 21, 22) Pin No Pin NO. Signal Name J5-E1 TXD0 1 TXD0 J5-E2 RXD0 2 RXD0 J5-D1 RTS0 3 RTS0 J5-D2 CTS0 4 CTS0 J5-E3 TRIGGER0 5 TRIGGER0 J5-E4 TXD1 6 TXD1 J5-E5 RXD1 7 RXD1 J5-D4 RTS1 8 RTS1 J5-D5 CTS1 9 CTS1 J5-D3 TRIGGER1 10 TRIGGER1 J5-E6 TXD2 11 TXD2 J5-E7 RXD2 12 RXD2 J5-D6 RTS2 13 RTS2 J5-D7 CTS2 14 CTS2 J5-E8 TRIGGER2 15 TRIGGER2 J5-E9 TXD3 16 TXD3 J5-E10 RXD3 17 RXD3 J5-D9 RTS3 18 RTS3 J5-D10 CTS3 19 CTS3 J5-D8 TRIGGER3 20 TRIGGER3 Active I/O H I L O Note ISH BCR1 ISH BCR2 L O R1CR L O R2CR L O J5-A1 TXD4 21 TXD4 J5-A2 RXD4 22 RXD4 J5-B1 RTS4 23 RTS4 J5-B2 CTS4 24 CTS4 J5-A3 TRIGGER4 25 TRIGGER4 J5-A4 CD5 26 CD5 J5-A6 RXD5 27 RXD5 J5-A5 TXD5 28 TXD5 J5-B7 DTR5 29 DTR5 J5-A7 DSR5 30 DSR5 J5-B5 RTS5 31 RTS5 J5-B6 CTS5 32 CTS5 J5-B4 CI5 33 CI5 34 GND O 35 GND O J5-A10 SHUTTER ON 36 SHUTTER ON J5-A9 LEDVCC 37 J5-B9 OPT TRIGGER L O CLI O Shutter On Signal LEDVCC O LED Vcc 38 GND O LED Vcc GND 39 OPT TRIGGER O Optical Trigger 40 GND CNN14 SLOT LUI L H O (To CNN Board CNN20) (Not Used) Pin No Pin NO. Signal Name Active I/O Note SLOT5 J5-E6 DC4 1 DC4 L O LED1-G (SM_AC/DC) J5-E7 DC5 2 DC5 L O LED1-O J5-E8 DC6 3 DC6 L O LED2-G J5-E9 DC7 4 DC7 L O LED2-O J5-E10 DC8 5 DC8 L O LED3-G J5-E11 DC9 6 DC9 L O LED3-O J5-E12 DC10 7 DC10 L O LED4-G J5-E13 DC11 8 DC11 L O LED4-O J5-E14 DC12 9 DC12 L O LED5-G J5-E15 DC13 10 DC13 L O LED5-O J4-D16 DC14 11 DC14 L O LED6-G J4-D17 DC15 12 DC15 L O LED6-O J4-D18 DC16 13 DC16 L O LED7-G J4-D19 DC17 14 DC17 L O LED7-O J4-D20 DC18 15 DC18 L O LED8-G J4-D21 DC19 16 DC19 L O LED8-O J4-D22 DC20 17 DC20 L O PULL-LED-G J4-D23 DC21 18 DC21 L O PULL-LED-O J4-D24 DC22 19 DC22 L O RUN-LED J4-D25 DC23 20 DC23 L O STAD-LED J5-D3 DC24 21 DC24 L O RERUN-LED J5-D4 DC25 22 DC25 L O STAD-LED-G J5-D5 DC26 23 DC26 L O STAD-LED-O J5-D6 DC27 24 DC27 L O DISK-LED-G J5-D7 DC28 25 DC28 L O DISK-LED-O J5-D8 DC29 26 DC29 L O SET-LED J5-C15 SENS30 27 SENS30 H I J5-C16 SENS31 28 SENS31 H I J5-C17 SENS32 29 SENS32 H I J5-C18 SENS33 30 SENS33 H I J5-D10 SENS34 31 SENS34 H I J5-D11 SENS35 32 SENS35 H I J5-D12 SENS36 33 SENS36 H I J5-D13 SENS37 34 SENS37 H I J5-D14 SENS38 35 SENS38 H I J4-E5 SENS39 36 SENS39 H I J4-E6 SENS40 37 SENS40 H I J4-E7 SENS41 38 SENS41 H I J4-E8 SENS42 39 SENS42 H I J4-E9 SENS43 40 SENS43 H I J4-E22 SENS44 41 SENS44 H I J4-E23 SENS45 42 SENS45 H I J4-E24 SENS46 43 SENS46 H I J4-E25 SENS47 44 SENS47 H I J5-E1 SENS48 45 SENS48 H I J5-E2 SENS49 46 SENS49 H I 47 GND O 48 GND O 49 GND O 50 GND O CNN15 Power Monitor Pin NO. Signal Name Code 1 +24V Monitor +24V 2 +24V GND +24V GND 3 +12V Monitor S+12V 4 +12V GND S+12V GND 5 +11.5V Monitor +11.5V +11.5V Monitor +11.5V GND CNN 16 Fan Pin NO. 1 2 Signal Name +12V Monitor Code S+12V 3 +12V Monitor S+12V GND ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064. All rights reserved. MECH LED The MECH LED board indicates sensor signal status for AC/DC, SMC and SM_AC/DC board input signals. The board is connected by three (3) cables to the AC/DC, SMC and SM_AC/DC Controller boards. MECH LED Board Reference LED No Board Artwork Function 1-1 RUHOME Reaction Carousel HOME sensor 2 RUMIR Reaction Carousel PITCH sensor 3 IAAHOME ICT rotational HOME 4 IAEHOME ICT Elevator HOME sensor 5 IAE MIR ICT Elevator PITCH sensor 6 STA1HOME Mixer 1 rotational HOME sensor 7 STA2HOME Mixer 2 rotational HOME sensor 8 R1CIHOME Reagent Supply Center 1 Inner Carousel HOME 9 R1CIPITCH Reagent Supply Center1 Inner Carousel PITCH 10 R1COHOME Reagent Supply Center 1 Outer carousel HOME LED No Board Artwork Function 2-1 R1COPITCH Reagent Supply Center 1 Outer Carousel PITCH sensor 2 R2CHOME Reagent Supply Center 2 Carousel HOME sensor 3 R2CPITCH Reagent Supply Center 2 Carousel PITCH sensor 4 SAAHOME Sample Pipettor rotational HOME sensor 5 SAAMIR Sample Pipettor rotational PITCH sensor 6 SAEHOME Sample Pipettor elevator HOME sensor 7 SAEMIR Sample Pipettor elevator PITCH sensor 8 SACLASH Sample Pipettor CRASH sensor 9 SPHOME Sample Syringe Drive HOME sensor 10 SPMIR Sample Syringe Drive PITCH sensor LED No Board Artwork Function 3-1 SPDLMT Sample Syringe Drive DOWN LIMIT sensor 2 R1PHOME Reagent 1 Syringe Drive HOME sensor 3 R1PMIR Reagent 1 Syringe Drive PITCH sensor 4 R1PDLMT Reagent 1 Syringe Drive DOWN LIMIT sensor 5 R2PHOME Reagent 2 Syringe Drive HOME sensor 6 R2PMIR Reagent 2 Syringe Drive PITCH sensor 7 R2PDLMT Reagent 2 Syringe Drive DOWN LIMIT sensor 8 R1AAHOME Reagent Pipettor 1 rotational HOME sensor 9 R1AAMIR Reagent Pipettor 1 rotational PITCH sensor 10 R1AEHOME Reagent Pipettor 1 Elevator HOME sensor LED No Board Artwork Function 4-1 R1AEMIR Reagent Pipettor 1 Elevator PITCH sensor 2 R1ACLASH Reagent Pipettor 1 CRASH sensor 3 R2AAHOME Reagent Pipettor 2 rotational HOME sensor 4 R2AAMIR Reagent Pipettor 2 rotational PITCH sensor 5 R2AEHOME Reagent Pipettor 2 Elevator HOME sensor 6 R2AEMIR Reagent Pipettor 2 PITCH sensor 7 R2ACLASH Reagent Pipettor 2 CRASH sensor 8 ISIHOME Sample Carousel HOME sensor 9 ISIPITCH Sample Carousel PITCH sensor 10 ISOHOME Inner Sampler Outer Home LED No Board Artwork Function 5-1 ISOPITCH Inner Sample Outer 1 Pitch 2 AHOME Auxiliary Stepping Motor Home 3 AMIR Auxiliary Stepping Motor 1 Round 4 SALVL Sample Pipettor Liquid Level Sensor 5 R1ALVL Reagent Pipettor 1 Liquid Level Sensor 6 R2ALVL Reagent Pipettor 2 Liquid Level Sensor 7 S Sample Pipettor Air Aspiration 8 9 10 LED No 6-1 Board Artwork WSUP Function Cuvette Wash Station UP sensor 2 WSDOWN Cuvette Wash Station DOWN sensor 3 WSDOWNOK Cuvette Wash Station DOWN-OK sensor 4 STEUP Mixer Elevator UP sensor 5 STEDOWN Mixer Elevator DOWN sensor 6 STEDOK Mixer Elevator DOWN-OK sensor 7 DPUP Wash Solution Pump UP sensor 8 DPDOWN Wash Solution Pump DOWN sensor 9 CWPUP Cuvette Wash UP sensor 10 CWPDOWN Cuvette Wash DOWN sensor LED No Board Artwork Function 7-1 PWPUP Inner Probe Wash Pump UP sensor 2 PWPDOWN Inner Probe Wash Pump DOWN sensor 3 TWPUP H.C. Waste Pump UP sensor 4 TWPDOWN H.C. Waste Pump DOWN sensor 5 ISPUP ICT Aspiration Pump UP sensor 6 ISPDOWN ICT Aspiration Pump DOWN sensor 7 IMPUP ICT Reference Solution Pump UP sensor 8 AC1UP Auxiliary UP Sensor 9 AC1DOWN Auxiliary DOWN Sensor 10 AC1OK Auxiliary OK Sensor LED No Board Artwork Function 8-1 AC2UP Auxiliary AC Motor 2 UP Sensor 2 AC2DOWN Auxiliary AC Motor 2 DOWN Sensor 3 AUXMUP Auxiliary DC Motor UP Sensor 4 AUXMDOWN Auxiliary DC Motor DOWN Sensor 5 BATHLVL Water Bath Float Switch OK 6 R1CCVROK Reagent Supply Center 1 Cover Sensor 7 R2CCVROK Reagent Supply Center 2 Cover Sensor (c4000 front cover sensor) 8 FFS1 Drain Tank L Float Switch 9 FFS2 Open Water Tank Upper Limit Float Switch 10 FFS3 Open Water Tank Level HIGH Float Switch LED No Board Artwork Function 9-1 FFS4 Open Water Tank Level LOW Float Switch 2 FFS5 Auxiliary Float Switch 3 ISCVROK Sample Carousel Cover Sensor (c4000 rear cover sensor) 4 DTF L.C. Waste Tank Level HIGH Float Switch 5 DALARM Degasser Vacuum Tank Abnormal Sensor Alarm 6 ISSW Inner Sampler Switch ON 7 ESMCC Empty Sensor MCC Warning Bulk Solution LOW Sensor, ICT Reference Solution 8 ESB14 Empty Sensor B14 Warning Bulk Solution LOW Sensor, Alkaline 9 ESB3 Empty Sensor B3 Warning Bulk Solution LOW Sensor, Acid 10 CUPSNSRI Reagent Bottle Sensor LED No Board Artwork Function 10-1 CUPSNSRO Inner Sampler Cup Sensor (OUT) ON 2 AUXSW1 Sample Carousel Access Indicator Switch (Pause) 3 AUXSW2 Sample Carousel Advance Indicator Switch (Scan) 4 ESMCCSS Empty Sensor MCCC S.S. (ICT Reference Solution Bulk Sensor Status) 5 ESB14SS Empty Sensor B14 S.S. (Alkaline Bulk Sensor Status) 6 ESB3SS Empty Sensor B3 S.S. (Acid Bulk Sensor Status) LED No Function (C16 Instrument) Board Artwork 11-1 SM1HOME Stirrer 1B Horizontal Movement Home 2 SM2HOME Stirrer 2B Horizontal Movement Home 3 SM3HOME Reagent 2 Outer Carousel Rotation Home 4 SM3PITCH Reagent 2 Outer Carousel Rotation Pitch 5 SM4HOME Sampler Pump B Line Up/Down Home 6 SM4PITCH Sampler Pump B Line Up/Down Pitch 7 SM4DLMT Sampler Pump B Line Up/Down Limit 8 SM5HOME Reagent 1B Pump Up/Down Home 9 SM5PITCH Reagent 1B Pump Up/Down Pitch 10 SM5DLMT Reagent 1B Pump Up/Down Limit LED No Board Artwork Function (C16 Instrument) 12-1 SM6HOME Reagent Pump 2B Motor Up/Down Home 2 SM6PITCH Reagent Pump 2B Motor Up/Down Pitch 3 SM6DLMT Reagent Pump 2B Motor Up/Down Limit 4 SM7HOME Reagent Arm 1B Horizontal Movement Home 5 SM7PITCH Reagent Arm 1B Horizontal Movement Pitch 6 SM8HOME Reagent Arm 1B Vertical Movement Home 7 SM8PITCH Reagent Arm 1B Vertical Movement Pitch 8 SM8CLASH Reagent Arm 1B Vertical Movement Clash 9 SM9HOME Reagent Arm 2B Horizontal Movement Home 10 SM9PITCH Reagent Arm 2B Horizontal Movement Pitch LED No Function (C16 Instrument) Board Artwork 13-1 SM10HOME Reagent Arm 2B Vertical Movement Home 2 SM10PITCH Reagent Arm 2B Vertical Movement Pitch 3 SM10CLASH Reagent Arm 2B Vertical Movement Clash 4 SM11HOME Not Used on C16000 5 SM11PITCH Not Used on C16000 6 SM11B Not Used on C16000 7 SM11F Not Used on C16000 8 SM12HOME High Concentration Waste Pump, Line B Home 9 SM12PITCH(OK) High Concentration Waste Pump, Line B Pitch (OK) 10 SM13HOME Sample Pump A Up/Down Motor Home LED No Board Artwork Function 14-1 SM13PITCH(OK) Sample Pump A Up/Down Motor Pitch (OK) 2 SM13DLMT Sample Pump A Up/Down Motor DLMT 3 AC1UP(HOME) Probe Wash Pump Line B Home 4 AC1DOWN(MOVE) Probe Wash Pump Line B Move 5 AC1S9 Probe Wash Pump Line B Sensor3 6 AC1OK Probe Wash Pump Line B Sensor4 7 SENS1 Stirrer 1A Cuvette 8 SENS2 Stirrer 2A Cuvette 9 SENS3 Stirrer 1B Cuvette 10 SENS4 Stirrer 2B Cuvette LED No Board Artwork Function 15-1 SENS5 IAVP-FB ISE Aspiration Pump Valve 2 SENS6 ESPV-FB External Sampling Probe Valve 3 SENS7 ERPV-FB External Reagent Probe Valve 4 SENS8 SWV-FB Stirrer Wash Valve 5 SENS9 WBSV-FB Water Bath Supply Pump Valve 6 SENS10 PWV-FB Pressure Water Valve 7 SENS11 CWV-FB Cuvette Wash Valve 8 SENS12 DTG1V-FB Detergent Switching Valve 9 SENS13 DTG2V-FB Detergent Backflow Prevention 10 SENS14 CDV-FB Cell Dryer Vacuum Pump Valve LED No Board Artwork Function 16-1 SENS15 ISPV-FB Internal Sampling Pump Value 2 SENS16 IR1PV-FB Internal Reagent 1 Probe Value 3 SENS17 IR2PV-FB Internal Reagent 2 Probe Pump 4 SENS18 ISPBV-FB Internal Sampling Line B Valve 5 SENS19 IR1BPV-FB Internal Reagent 1B Probe Valve 6 SENS20 IR2BPV-FB Internal Reagent 2B Probe Valve 7 SENS21 Buzzer-FB 8 SENS22 IREFCUP-FB IREF Cup Sensor 9 SENS23 AUXDC1-FB Spare 10 SENS24 VP2-FB Vacuum Pump 2 Aspirate LED No Board Artwork Function 17-1 SENS25 VP1-FB Vacuum Pump 1 Dry Cuvette 2 SENS26 TWMP-FB DI Water Supply Pump 1 3 SENS27 WPMP-FP Wash Pools Magnet Pump 4 SENS28 MC-FB Freezer Power Control Relay 5 SENS29 AUXAC3-ON-FB 6 SENS30 AUXSENS1 7 SENS31 AUXSENS2 8 SENS32 AUXSENS3 9 SENS33 AUXSENS4 10 SENS34 AUXSENS5 LED No Board Artwork Function 18-1 SENS35 AUXSENS6 2 SENS36 AUXSENS7 3 SENS37 AUXSENS8 4 SENS38 AUXSENS9 5 SENS39 AUXSENS10 6 SENS40 AUXSENS11 7 SENS41 AUXSENS12 8 SENS42 AUXSENS13 9 SENS43 AUXSENS14 10 SENS44 AUXSENS15 LED No Board Artwork Function 19-1 SENS45 AUXSENS16 2 SENS46 AUXSENS17 3 SENS47 AUXSENS18 4 SENS48 AUXSENS19 5 SENS49 AUXSENS20 6 ---------- ARCHITECT c4000 Service and Support Manual (Version 204733-112) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. New Style MECH LED The New Style MECH LED board indicates sensor signal status for AC/DC, SMC and SM_AC/DC board input signals. The board is connected by three (3) cables to the AC/DC, SMC and SM_AC/DC Controller boards. MECH LED Board Reference LED No Board Artwork Function D1 RUHOME Reaction Carousel HOME sensor D2 RUMIR Reaction Carousel PITCH sensor D3 IAAHOME ICT rotational HOME D4 IAEHOME ICT Elevator HOME sensor D5 IAE MIR ICT Elevator PITCH sensor D6 STA1HOME Mixer 1 rotational HOME sensor D7 STA2HOME Mixer 2 rotational HOME sensor D8 R1CIHOME Reagent Supply Center 1 Inner Carousel HOME D9 R1CIPITCH Reagent Supply Center1 Inner Carousel PITCH D10 R1COHOME Reagent Supply Center 1 Outer carousel HOME D11 R1COPITCH Reagent Supply Center 1 Outer Carousel PITCH sensor D12 R2CHOME Reagent Supply Center 2 Carousel HOME sensor D13 R2CPITCH Reagent Supply Center 2 Carousel PITCH sensor D14 SAAHOME Sample Pipettor rotational HOME sensor D15 SAAMIR Sample Pipettor rotational PITCH sensor D16 SAEHOME Sample Pipettor elevator HOME sensor D17 SAEMIR Sample Pipettor elevator PITCH sensor D18 SACLASH Sample Pipettor CRASH sensor D19 SPHOME Sample Syringe Drive HOME sensor D20 SPMIR Sample Syringe Drive PITCH sensor D21 SPDLMT Sample Syringe Drive DOWN LIMIT sensor D22 R1PHOME Reagent 1 Syringe Drive HOME sensor D23 R1PMIR Reagent 1 Syringe Drive PITCH sensor D24 R1PDLMT Reagent 1 Syringe Drive DOWN LIMIT sensor D25 R2PHOME Reagent 2 Syringe Drive HOME sensor D26 R2PMIR Reagent 2 Syringe Drive PITCH sensor D27 R2PDLMT Reagent 2 Syringe Drive DOWN LIMIT sensor D28 R1AAHOME Reagent Pipettor 1 rotational HOME sensor D29 R1AAMIR Reagent Pipettor 1 rotational PITCH sensor D30 R1AEHOME Reagent Pipettor 1 Elevator HOME sensor D31 R1AEMIR Reagent Pipettor 1 Elevator PITCH sensor D32 R1ACLASH Reagent Pipettor 1 CRASH sensor D33 R2AAHOME Reagent Pipettor 2 rotational HOME sensor D34 R2AAMIR Reagent Pipettor 2 rotational PITCH sensor D35 R2AEHOME Reagent Pipettor 2 Elevator HOME sensor D36 R2AEMIR Reagent Pipettor 2 PITCH sensor D37 R2ACLASH Reagent Pipettor 2 CRASH sensor D38 ISIHOME Sample Carousel HOME sensor D39 ISIPITCH Sample Carousel PITCH sensor D40 ISOHOME Inner Sampler Outer Home D41 ISOPITCH Inner Sample Outer 1 Pitch D42 AHOME Auxiliary Stepping Motor Home D43 AMIR Auxiliary Stepping Motor 1 Round D44 SALVL Sample Pipettor Liquid Level Sensor D45 R1ALVL Reagent Pipettor 1 Liquid Level Sensor D46 R2ALVL Reagent Pipettor 2 Liquid Level Sensor D47 S Sample Pipettor Air Aspiration D48 WSUP Cuvette Wash Station UP sensor D49 WSDOWN Cuvette Wash Station DOWN sensor D50 WSDOWNOK Cuvette Wash Station DOWN-OK sensor D51 STEUP Mixer Elevator UP sensor D52 STEDOWN Mixer Elevator DOWN sensor D53 STEDOK Mixer Elevator DOWN-OK sensor D54 DPUP Wash Solution Pump UP sensor D55 DPDOWN Wash Solution Pump DOWN sensor D56 CWPUP Cuvette Wash UP sensor D57 CWPDOWN Cuvette Wash DOWN sensor D58 PWPUP Inner Probe Wash Pump UP sensor D59 PWPDOWN Inner Probe Wash Pump DOWN sensor D60 TWPUP H.C. Waste Pump UP sensor D61 TWPDOWN H.C. Waste Pump DOWN sensor D62 ISPUP ICT Aspiration Pump UP sensor D63 ISPDOWN ICT Aspiration Pump DOWN sensor D64 IMPUP ICT Reference Solution Pump UP sensor D65 AC1UP Auxiliary UP Sensor D66 AC1DOWN Auxiliary DOWN Sensor D67 AC1OK Auxiliary OK Sensor D68 AC2UP Auxiliary AC Motor 2 UP Sensor D69 AC2DOWN Auxiliary AC Motor 2 DOWN Sensor D70 AUXMUP Auxiliary DC Motor UP Sensor D71 AUXMDOWN Auxiliary DC Motor DOWN Sensor D72 BATHLVL Water Bath Float Switch OK D73 R1CCVROK Reagent Supply Center 1 Cover Sensor D74 R2CCVROK Reagent Supply Center 2 Cover Sensor (c4000 front cover sensor) D75 FFS1 Drain Tank L Float Switch D76 FFS2 Open Water Tank Upper Limit Float Switch D77 FFS3 Open Water Tank Level HIGH Float Switch D78 FFS4 Open Water Tank Level LOW Float Switch D79 FFS5 Auxiliary Float Switch D80 ISCVROK Sample Carousel Cover Sensor (c4000 rear cover sensor) D81 DTF L.C. Waste Tank Level HIGH Float Switch D82 DALARM Degasser Vacuum Tank Abnormal Sensor Alarm D83 ISSW Inner Sampler Switch ON D84 ESMCC Empty Sensor MCC Warning Bulk Solution LOW Sensor, ICT Reference Solution D85 ESB14 Empty Sensor B14 Warning Bulk Solution LOW Sensor, Alkaline D86 ESB3 Empty Sensor B3 Warning Bulk Solution LOW Sensor, Acid D87 CUPSNSRI Reagent Bottle Sensor D88 CUPSNSRO Inner Sampler Cup Sensor (OUT) ON D89 AUXSW1 Sample Carousel Access Indicator Switch (Pause) D90 AUXSW2 Sample Carousel Advance Indicator Switch (Scan) D91 ESMCCSS Empty Sensor MCCC S.S. (ICT Reference Solution Bulk Sensor Status) D92 ESB14SS Empty Sensor B14 S.S. (Alkaline Bulk Sensor Status) D93 ESB3SS Empty Sensor B3 S.S. (Acid Bulk Sensor Status) D94 SM1HOME Stirrer 1B Horizontal Movement Home D95 SM2HOME Stirrer 2B Horizontal Movement Home D96 SM3HOME Reagent 2 Outer Carousel Rotation Home D97 SM3PITCH Reagent 2 Outer Carousel Rotation Pitch D98 SM4HOME Sampler Pump B Line Up/Down Home D99 SM4PITCH Sampler Pump B Line Up/Down Pitch D100 SM4DLMT Sampler Pump B Line Up/Down Limit D101 SM5HOME Reagent 1B Pump Up/Down Home D102 SM5PITCH Reagent 1B Pump Up/Down Pitch D103 SM5DLMT Reagent 1B Pump Up/Down Limit D104 SM6HOME Reagent Pump 2B Motor Up/Down Home D105 SM6PITCH Reagent Pump 2B Motor Up/Down Pitch D106 SM6DLMT Reagent Pump 2B Motor Up/Down Limit D107 SM7HOME Reagent Arm 1B Horizontal Movement Home D108 SM7PITCH Reagent Arm 1B Horizontal Movement Pitch D109 SM8HOME Reagent Arm 1B Vertical Movement Home D110 SM8PITCH Reagent Arm 1B Vertical Movement Pitch D111 SM8CLASH Reagent Arm 1B Vertical Movement Clash D112 SM9HOME Reagent Arm 2B Horizontal Movement Home D113 SM9PITCH Reagent Arm 2B Horizontal Movement Pitch D114 SM10HOME Reagent Arm 2B Vertical Movement Home D115 SM10PITCH Reagent Arm 2B Vertical Movement Pitch D116 SM10CLASH Reagent Arm 2B Vertical Movement Clash D117 SM11HOME Not Used on C16000 D118 SM11PITCH Not Used on C16000 D119 SM11B Not Used on C16000 D120 SM11F Not Used on C16000 D121 SM12HOME High Concentration Waste Pump, Line B Home D122 SM12PITCH(OK) High Concentration Waste Pump, Line B Pitch (OK) D123 SM13HOME Sample Pump A Up/Down Motor Home D124 SM13PITCH(OK) Sample Pump A Up/Down Motor Pitch (OK) D125 SM13DLMT Sample Pump A Up/Down Motor DLMT D126 AC1UP(HOME) Probe Wash Pump Line B Home D127 AC1DOWN(MOVE) Probe Wash Pump Line B Move D128 AC1S9 Probe Wash Pump Line B Sensor3 D129 AC1OK Probe Wash Pump Line B Sensor4 D130 SENS1 Stirrer 1A Cuvette D131 SENS2 Stirrer 2A Cuvette D132 SENS3 Stirrer 1B Cuvette D133 SENS4 Stirrer 2B Cuvette D134 SENS5 IAVP-FB ISE Aspiration Pump Valve D135 SENS6 ESPV-FB External Sampling Probe Valve D136 SENS7 ERPV-FB External Reagent Probe Valve D137 SENS8 SWV-FB Stirrer Wash Valve D138 SENS9 WBSV-FB Water Bath Supply Pump Valve D139 SENS10 PWV-FB Pressure Water Valve D140 SENS11 CWV-FB Cuvette Wash Valve D141 SENS12 DTG1V-FB Detergent Switching Valve D142 SENS13 DTG2V-FB Detergent Backflow Prevention D143 SENS14 CDV-FB Cell Dryer Vacuum Pump Valve D144 SENS15 ISPV-FB Internal Sampling Pump Value D145 SENS16 IR1PV-FB Internal Reagent 1 Probe Value D146 SENS17 IR2PV-FB Internal Reagent 2 Probe Pump D147 SENS18 ISPBV-FB Internal Sampling Line B Valve D148 SENS19 IR1BPV-FB Internal Reagent 1B Probe Valve D149 SENS20 IR2BPV-FB Internal Reagent 2B Probe Valve D150 SENS21 Buzzer-FB D151 SENS22 IREFCUP-FB IREF Cup Sensor D152 SENS23 AUXDC1-FB Spare D153 SENS24 VP2-FB Vacuum Pump 2 Aspirate D154 SENS25 VP1-FB Vacuum Pump 1 Dry Cuvette D155 SENS26 TWMP-FB DI Water Supply Pump 1 D156 SENS27 WPMP-FP Wash Pools Magnet Pump D157 SENS28 MC-FB Freezer Power Control Relay D158 SENS29 AUXAC3-ON-FB D159 SENS30 AUXSENS1 D160 SENS31 AUXSENS2 D161 SENS32 AUXSENS3 D162 SENS33 AUXSENS4 D163 SENS34 AUXSENS5 D164 SENS35 AUXSENS6 D165 SENS36 AUXSENS7 D166 SENS37 AUXSENS8 D167 SENS38 AUXSENS9 D168 SENS39 AUXSENS10 D169 SENS40 AUXSENS11 D170 SENS41 AUXSENS12 D171 SENS42 AUXSENS13 D172 SENS43 AUXSENS14 D173 SENS44 AUXSENS15 D174 SENS45 AUXSENS16 D175 SENS46 AUXSENS17 D176 SENS47 AUXSENS18 D177 SENS48 AUXSENS19 D178 SENS49 AUXSENS20 D179 LVL1OK Reagent Pipettpr 1B Liquid Level Sensor D180 LVL2OK Reagent Pipettpr 2B Liquid Level Sensor ARCHITECT c4000 Service and Support Manual (Version 204733-122) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. PM The PM subsystem detects abnormalities, such as; clots (probe obstructions), air bubbles in patient sample, air bubbles and large amounts of foam in reagent bottles during aspiration and dispense operations. Wash pressure is monitored during internal probe washing to determine if the pressure sensed is below or above pressure thresholds (referenced in the c4000 PM Board firmware). PM Board PM Board (side view) PM Board (end view) Rocker Switch Settings Switch Settings 1 OFF 2 OFF 3 OFF 4 OFF 5 ON 6 OFF 7 ON 8 OFF PM Block Diagram ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. SMC SMC Board The SMC-C board controls the 5-phase and 2 phase stepping-motors S1 Switch Settings Switch S1-1 Setting (*Default) Function Exhibition Mode (see P-644 Demonstration Mode for more information) ON Normal Run Mode OFF* S1-2 UNUSED OFF* S1-3 Factory Use Only OFF* S1-4 Do not initialize 2D BCR at Power On ON Initialize 2D BCR at Power On OFF* One ACD/ALK Valve ON Two ACD/ALK Valves OFF* S1-6 Instrument Type (Factory use only) OFF* S1-7 Software Install Mode (Factory use only) OFF* S1-8 Software Boot Up Mode (Factory use only) OFF* S1-5 SMC Monitoring LEDs LED # LED 1 Motor Name/Code Reaction Carousel Motor (RUM) LED # LED 11 Motor Name/Code Reagent Syringe1 Drive Motor (R1PM) LED 2 ICT Rotation Motor (IAAM) LED12 Reagent Syringe 2 Drive Motor (R2PM) LED 3 ICT Elevator Motor (IAEM) LED 13 Reagent 1 Pipettor Rotation Motor (R1AAM) LED 4 Mixer 1 Rotation Motor (STAM1) LED 14 Reagent 2 Pipettor Rotation Motor (R2AAM) LED 5 Mixer 2 Rotation Motor (STAM2) LED 15 Sample Carousel Rotation Motor (ISIM) LED 6 Reagent Supply Center 1 Inner Carousel Rotation Motor ((R1CIM) LED 16 Inner Sampler ISOM LED 7 Reagent Supply Center 1 Outer Carousel Rotation Motor (R1COM) LED 17 Sample Pipettor Elevator Motor (SAEM) LED 8 Reagent Supply Center 2 Carousel Rotation Motor (S2CM) LED 18 Reagent 1 Pipettor Elevator Motor (R1AEM) LED 9 Sample Pipettor Rotation Motor (SAAM) LED 19 Reagent 2 Pipettor Elevator Motor (R2AEM) LED 10 Sample Syringe Drive Motor (SPM) LED 20 (AUXSM) SMC Block Diagram Schematic Diagram Link SMC Block Diagram 7A_9010a.pdf 1 PMC (Pulse Motor Controller) There are a total of 20 LSI (two types) pulse motor controllers mounted on the SMC Board. 2 Control EPLD The control ELPD (Electrically Programmable Logic Device) includes a majority logic circuit for sensor input, a sensor, a status read register, and the counter circuit for the pitch sensor. This circuit uses a PLD. 3 PMC monitoring LED These LED's monitor the operating status of the PMC's. The color is red. 4 Debugging LED's and Switches 8-bit LED and 8-bit DIPSW used during development. The color is red. 5 Power ON LED Indicator LED is ON when +5 volts DC is supplied. The color is green. 6 Compact PCI Interface Circuit This interface controls the interface bus with the host CPU. This circuit uses a PLD. List of SMC Stepper Motor Controls (SMC) Stepper Motor No. 1 2 Name Reaction Carousel Motor ICT Rotation Motor Code RUM IAAM Sensor Phase 5f 2f Signal Name Signal Name Description RUCW RUHOME Detects the home position RUCCW RUM1R Detects a single motor rotation IAAINA IAAHOME Detects the home position IAEHOME IAEM1R Detects the home position Detects a single motor rotation STA1HOME Detects the home position IAATDA IAAINB IAATDB IAAHOLD 3 ICT Elevator Motor IAEM 2f IAEINA IAETDA IAEINB IAETDB IAEHOLD 4 Mixer 1 Rotation Motor STAM1 2f STA1NA STA1TDA STA1NB STA1TDB STA1HOLD 5 Mixer 2 Rotation Motor STAM2 2f STA2INA STA2HOME Detects the home position R1CICW R1CIHOME Detects the home position R1CICCW R1CIPITCH Detects a single motor rotation R1COCW R1COHOME Detects the home position R1COCCW R1COPITCH Detects a single motor rotation R2CCW R2CHOME Detects the home position R2CCW R2CPITCH Detects a single motor rotation SAACW SAAHOME Detects the home position SAACCW SAAM1R Detects a single motor rotation SAECW SAEHOME Detects the home position SAECCW SAEM1R Detects a single motor rotation SACLASH Detects an obstacle SALVL Detects the liquid surface SPINA SPHOME Detects the home position SPTDA SPM1R Detects a single motor rotation SPINB SPDLMT Detects the DOWN limit R1PINA R1PHOME Detects the home position R1PTDA R1PM1R Detects a single motor rotation R1PINB R1PDLMT Detects the DOWN limit R2PINA R2PHOME Detects the home position R2PTDA R2PM1R Detects a single motor rotation R2PINB R2PDLMT Detects the DOWN limit R1AAINA R1AAHOME Detects the home position R1AATDA R1AAM1R Detects a single motor rotation R1AEINA R1AEHOME Detects the home position R1AETDA R1AEM1R Detects a single motor rotation R1AEINB R1ACLASH Detects an obstacle R1AETDB R1ALVL Detects the liquid surface STA2TDA STA2INB STA2TDB STA2HOLD 6 7 8 9 10 11 R-1 Supply C. Inner Motor R-1 Supply C. Outer Motor R-2 Supply C. Motor Sample Pipettor Rotation Motor Sample Pipettor Elevator Motor Sample Syringe Drive Motor R1CIM R1COM R2CM SAAM SAEM SPM 5f 5f 5f 5f 5f 2f SPTDB SPHOLD 12 Reagent 1 Syringe Drive Motor R1PM 2f R1PTDB R1PHOLD 13 Reagent 2 Syringe Drive Motor R2PM 2f R2TDB R2PHOLD 14 Reagent 1 Pipettor rotation motor R1AAM 2f R1AAINB R1AATDB R1AAHOLD 15 Reagent 1 Pipettor Elevator Motor R1AEM 2f R1AEHOLD 16 Reagent 2 Pipettor rotation motor R2AAM 2f R2AAINA R2AAHOME Detects the home position R2AATDA R2AAM1R Detects a single motor rotation R2AEINA R2AEHOME Detects the home position R2AETDA R2AEM1R Detects a single motor rotation R2AEINB R2ACLASH Detects an obstacle R2AETDB R2LVL Detects the liquid surface ISIINA ISIHOME Detects the home position ISITDA ISIPITCH Detects a single motor rotation ISOINA ISOHOME Detects the home position ISOTDA ISOPITCH Detects a single motor rotation AUXINA AHOME Detects the home position AUXTDA AM1R Detects a single motor rotation R2AAINB R2AATDB R2AAHOLD 17 Reagent 2 Pipettor Elevator Motor R2AEM 2f R2AEHOLD 18 Sample Carousel Rotation Motor (not for C4000) ISIM 2f ISIINB ISITDB ISIHOLD 19 Inner Sampler (not for C4000) ISOM 2f ISOINB ISOTDB ISOHOLD 20 Inner Sampler (not for C4000) AUXSM 2f AUXINB AUXTDB AUXHOLD SMC Front Panel Connector CN1 (True value = Low) Pin No. Signal Name Code 1 Home position sensor for the reaction carousel motor RUHOME 2 Single motor rotation sensor (pitch) for the reaction carousel motor RUM1R 3 Home position sensor the ICT rotation motor IAAHOME 4 Home position sensor for the ICT elevator motor IAEHOME 5 Single motor rotation sensor (pitch) for the ICT elevator motor IAEM1R 6 Home position sensor for Mixer 1 rotation motor STA1HOME 7 Home position sensor for Mixer 2 rotation motor STA2HOME 8 Home position sensor for reagent supply center 1 inner carousel motor R1CIHOME 9 Single motor rotation sensor (pitch) for reagent supply center 1 inner carousel motor R1CIPITCH 10 Home position sensor for the reagent supply center 1 outer carousel motor R1COHOME 11 Single motor rotation sensor (pitch) for reagent supply center 1 outer carousel motor R1COPITCH 12 Home position sensor for reagent supply center 2 carousel motor R2CHOME 13 Single motor rotation sensor (pitch) for reagent supply center 2 carousel R2CPITCH 14 Home position sensor for the sample pipettor rotation motor SAAHOME 15 Single motor rotation sensor (pitch) for the sample pipettor rotation motor SAAM1R 16 Home position sensor for the sample pipettor elevator motor SAEHOME 17 Single motor rotation sensor (pitch) for the sample pipettor elevator motor SAEM1R 18 Crash sensor for the sample pipettor SACLASH 19 Home position sensor for the sample syringe drive motor SPHOME 20 Single motor rotation sensor (pitch) for the sample syringe drive motor SPM1R 21 DOWN limit sensor for the sample syringe drive motor SPDLMT 22 Home position sensor for the reagent 1 syringe drive motor R1PHOME 23 Single motor rotation sensor (pitch) for the reagent 1 syringe drive motor R1PM1R 24 DOWN limit sensor for the reagent 1 syringe drive motor R1PDLMT 25 Home position sensor for the reagent 2 syringe drive motor R2PHOME 26 Single motor rotation sensor (pitch) for the reagent 2 syringe drive motor R2PM1R 27 DOWN limit sensor for the reagent 2 syringe drive motor R2PDLMT 28 Home position sensor for the reagent 1 pipettor rotation motor R1AAHOME 29 Single motor rotation sensor (pitch) for the reagent 1 pipettor rotation motor R1AAM1R 30 Home position sensor for the reagent pipettor elevator motor R1AEHOME 31 Single motor rotation sensor (pitch) for the reagent 1 pipettor elevator motor R1AEM1R 32 Crash sensor for the reagent 1 pipettor R1ACLASH 33 Home position sensor for the reagent 2 pipettor rotation motor R2AAHOME 34 Single motor rotation sensor (pitch) for the reagent 2 pipettor rotation motor R2AAM1R 35 Home position sensor for the reagent 2 pipettor elevator motor R2AEHOME 36 Single motor rotation sensor (pitch) for the reagent 2 pipettor elevator motor R2AEM1R 37 Crash sensor for the reagent 2 pipettor R2ACLASH 38 Home position sensor for the sample carousel rotation motor ISIHOME 39 Single motor rotation sensor (pitch) for the sample carousel rotation motor ISIPITCH 40 ISOHOME 41 ISOPITCH 42 AHOME 43 AM1R 44 LLS for the sample pipettor SALVL 45 LLS for the reagent 1 pipettor R1ALVL 46 LLS for the reagent 2 pipettor R2ALVL 47 Sample probe liquid surface non-contact status 48 blank 49 Power supply +5 VDC 50 Power supply +5 VDC 51 Power supply +5 VDC 52 Power supply +5 VDC SMC Controller Motor Drive Signals and Sensor Signals Pin A B C D E F 22 SALVLOK R1ALVL R1AAUTOZERO R2ALVLOK GND 21 VCC (SALVLOK) VCC (R1ALVL) VCC (R1AAUTOZERO) VCC (R2ALVLOK) GND J5 Connector 20 SALVL SAAUTOZERO R1ALVLOK R2ALVL R2AAUTOZERO GND 19 VCC (SALVL) VCC (SAAUTOZERO) VCC (R1ALVLOK) VCC (R2ALVL) VCC (R2AAUTOZERO) GND 18 GND 17 ACW R1PTDA SSTA2HOLD GND 16 ACCW R1PINB ISIINA GND 15 R1AEINA R1PTDB ISITDA GND 14 R1AETDA R1PHOLD ISIINB GND 13 R1AINB R2PINA ISITDB GND 12 R1AETDB R2PTDA ISIHOLD GND 11 R1AEHOLD R2PINB ISOINA GND 10 R1AAINA R2PTDB ISOTDA GND 9 R1AATDA R2PHOLD ISOINB GND 8 R1AAINB IAEINA ISOTDB GND 7 R1AATDB IAETDA ISOHOLD GND 6 R1AAHOLD +5V IAEINB +5V AUXINA GND 5 R2AEINA +5V IAETDB +5V AUXTDA GND 4 RWAETDA +5V IAEHOLD +5V AUXINB GND 3 R2AEINB GND IAAINA GND AUXTDB GND 2 R2AETDB GND IAATDA GND AUXHOLD GND 1 R2AEHOLD GND IAAINB GND RUCW GND 25 R2AAINA IAATDB RUCCW GND 24 R2AATDA IAAINB R1C1CW GND 23 R2AAINB STA1INA R1C1CCW GND 22 R2AATDB STA1TDA R1COCW GND 21 R2AAHOLD SSTA1INB R1COCCW GND 20 SPINA STA1TDB R2CCW GND 19 SPTDA STA1HOLD R2CCCW GND 18 SPINB STA2INA SAACW GND 17 SPTDB STA2TDA SAACCW GND 16 SPHOLD STA2INB SAECW GND 15 R1PINA STA2TDB SAECCW GND 14 KEY AREA 13 12 11 ISOPITCH AHOME AM1R 10 R2ACLASH ISIHOME ISIPITCH ISOHOME GND 9 R2AAHOME R2AAM1R R2AEHOME R2AEM1R GND 8 R1AAM1R R1AEHOME R1AEM1R R1ACLASH GND 7 R2PHOME R2PM1R R2PDLMT R1AAHOME SALVLGU GND 6 SPDLMT R1PHOME R1PM1R R1PDLMT SALVLGUV(+5v0 GND 5 SAEM1R SACLASH SPHOME SPM1R ISOPOS GND 4 R2CPITCH SAAHOME SAAM1R SAEHOME ISIPOS GND 3 R1C1PITCH R1COHOME R1COPITCH R2CHOME R2CPOS GND 2 IAEM1R STA1HOME STA2HOME R1CIHOME R1COPOS GND J4 Connector 1 RUHOME RUM1R IAAHOME IAEHOME R1CIPPOS GND ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. SM AC/DC SM AC/DC Controller PCB The SM AC/DC Controller Board controls stepping motors, AC motors and DC valves. The SM AC/DC Controller Board controls the reagent loading motors and associated sensors for the optional Continuos Reagent Access (CRA) function. SM AC/DC Controller Board SM AC/DC PCB Edge SM AC/DC Block Diagram Main Functional Areas 1 PMC (Pulse Motor Controller) There are two LSI (Large Scale Integration) pulse motor controller types mounted on the SM AC/DC Board. 2 FPGA Controller The FPGA (Field Programmable Gate Array) controller consists of AC motor, DC valve drive circuit, the majority decision logic for sensor input, sensor, lead register of status, and counter circuitry for PITCH sensors. 3 SM Status LED The stepper motor LED turns red when the Motor Control LSI (Large Scale Integration) is busy. 4 Debugging LED and Switch 16-bit DIP SW for debugging. The color of the LED is red. 5 Power ON LED The Power Indicator LED is green when the +5V is on. 6 PCI Interface This interface controls the interface bus with the host CPU. This circuit uses a FPGA. 7 RS232C Interface Provides for communication with a barcode scanner. Stepping (Pulse) Motor Control No. Name Phase Motor Signal Name Sensor Signal Name Description 1 Stirrer 1B Horizontal Movement (STR1BH) 2f SM1INA SM1TDA SM1INB SM1TDB SM1HOLD SM1HOME Detects the home position 2 Stirrer 2B Horizontal Movement (STR2BH) 2f SM2INA SM2TDA SM2INB SM2TDB SM2HOLD SM2HOME Detects the home position 3 Reagent 2 Outer Carousel (R2OC) 5f SM3CW SM3CCW SM3HOME SM3PITCH Detects the home position Detects a single motor rotation 4 Sampler Pump Line B (SPB) 2f SM4INA SM4TDA SM4INB SM4TDB SM4HOLD SM4HOME SM4PITCH SM4DLMT Detects the home position Detects a single motor rotation Detects the DOWN limit 5 Reagent 1B Pump (R1BP) 2f SM5INA SM5TDA SM5INB SM5TDB SM5HOLD SM5HOME SM5PITCH SM5DLMT Detects the home position Detects a single motor rotation Detect the DOWN limit 6 Reagent 2B Pump (R2BP) 2f SM6INA SM6TDA SM6INB SM6TDB SM6HOLD SM6HOME SM6PITCH SM6DLMT Detects the home position Detects a single motor rotation Detect the DOWN limit 7 Reagent 1B Arm Horizontal Movement (R1ABH) 2f SM7INA SM7TDA SM7INB SM7TDB SM7HOLD SM7HOME SM7PITCH Detects the home position Detects a single motor rotation 8 Reagent 1B Arm Vertical Movement (R1ABZ) 2f SM8INA SM8TDA SM8INB SM8TDB SM8HOLD SM8HOME SM8PITCH SM8CLASH LVL1 Detects the home position Detects a single motor rotation Detects an obstacle Detects the liquid surface 9 Reagent 2B Arm Horizontal Movement (R2ABH) 2f SM9INA SM9TDA SM9INB SM9HOME SM9PITCH Detects the home position Detects a single motor rotation SM9TDB SM9HOLD 10 Reagent 2B Arm Vertical Movement (R2ABZ) SM10INA SM10TDA SM10INB SM10TDB SM10HOLD SM10HOME SM10PITCH SM10CLASH LVL2 Detects the home position Detects a single motor rotation Detects an obstacle Detects the liquid surface 12 Waste Pump for High Conc waste, line B (PPBHC) 2f SM12INA SM12TDA SM12INB SM12TDB SM12HOLD SM12HOME SM12PITCH Detects the home position Detects a single motor rotation 13 Sampler Pump (SP) (CW/CCW pulse) 5f SM13HOME SM13PITCH SM13DLMT Detects the home position Detects a single motor rotation Detect the DOWN limit 1 Probe Wash Pump line B (PWBP) PWBP_UP PWBP_DOWN PWBP_BK Note The AC1 motor is controlled CW/CCW by commands from the CPU. AC Motor Control No. DC Control List No. Device/Signal Name Control Signal Code 1 Internal Sampling Probe line B Valve (ISPBV) ON/OFF DC1 Valve 2 Internal Reagent 1B Probe Valve (IR1BPV) ON/OFF DC2 Valve 3 Internal Reagent 2B Probe Valve (IR2BPV) ON/OFF DC3 Valve 4 LED1-G ON/OFF DC4 LED 5 LED1-O ON/OFF DC5 LED 6 LED2-G ON/OFF DC6 LED 7 LED2-O ON/OFF DC7 LED 8 LED3-G ON/OFF DC8 LED 9 LED3-O ON/OFF DC9 LED 10 LED4-G ON/OFF DC10 LED 11 LED4-O ON/OFF DC11 LED 12 LED5-G ON/OFF DC12 LED 13 LED5-O ON/OFF DC13 LED 14 LED6-G ON/OFF DC14 LED 15 LED6-O ON/OFF DC15 LED 16 LED7-G ON/OFF DC16 LED 17 LED7-O ON/OFF DC17 LED 18 LED8-G ON/OFF DC18 LED 19 LED8-O ON/OFF DC19 LED 20 PULL-LED-G ON/OFF DC20 LED 21 PULL-LED-O ON/OFF DC21 LED 22 RUN-LED ON/OFF DC22 LED 23 STAD-LED ON/OFF DC23 LED 24 RERUN-LED ON/OFF DC24 LED 25 STAD-LED-G ON/OFF DC25 LED 26 STAD-LED-O ON/OFF DC26 LED 27 DISK-LED-G ON/OFF DC27 LED 28 DISK-LED-O ON/OFF DC28 LED 29 SET-LED ON/OFF DC29 LED 30 1(TRIGGER) ON/OFF DC30 BCRT 31 2(TRIGGER) ON/OFF DC31 BCRT ON/OFF DC32 32 Individual Sensor Control No. Device or Signal Name Control SIgnal Code Sensor 1 STR1A CUV Read SENS-1 STR1H-CUV 2 STR2A CUV Read SENS-2 STR2H-CUV 3 STR1B CUV Read SENS3 STR1BH-CUV 4 STR2B CUV Read SENS4 STR2BH-CUV 5 ISE Aspir Pump Valve Read SENS5 IAVP-FB +12V H Active 6 DI Water Valve 1 (S-Outer) Read SENS6 ESPV-FB +12V H Active 7 DI Water Valve 2 (R-Outer) Read SENS7 ERPV-FB +12V H Active 8 DI Water Valve 3 (STR) Read SENS8 SWV-FB +12V H Active 9 DI Water Valve 4 (Water Bath) Read SENS9 WBSV-FB +12V H Active 10 DI Water Valve 5 (Pressurized) Read SENS10 PWV-FB +12V H Active 11 Cell Wash Valve Read SENS11 CWV-FB +12V H Active 12 Alkaline & Acid Valve 1 Read SENS12 DTG1V-FB +12V H Active 13 Alkaline & Acid Valve 2 Read SENS13 DTG2V-FB +12V H Active 14 Cell Dryer Vac Pump Valve Read SENS14 CDV-FB +12V H Active 15 Sampling Pump Valve Read SENS15 ISPV-FB +12V H Active 16 Reagent Pump 1A Valve Read SENS16 IR1PV-FB +12V H Active 17 Reagent Pump 2A Valve Read SENS17 IR1PV-FB +12V H Active 18 Sampling Pump B Valve Read SENS18 ISPBV-FB +12V H Active 19 Reagent Pump 1B Valve Read SENS19 +12V H Active IRBPV-FB 20 Reagent Pump 2B Valve Read SENS20 IRBPV-FB +12V H Active 21 Buzzer Read SENS21 BUZZER-FB +12V H Active 22 IRef Cup Sensor Read SENS22 IREFCUP-FB +12V H Active 23 Spare Read SENS23 AUXDC1-FB +12V H Active 24 Low Conc Vac Pump AB Read SENS24 VP2-FB +5V L Active 25 Cell Dryer Vac Pump AB Read SENS25 VP1-FB +5V L Active 26 DI Water Supply Pump 1 (Water Pump) Read SENS26 TWMP-FB +5V L Active 27 DI Water Supply Pump 1 (Outer Wash) Read SENS27 WPMP-FB +5V L Active 28 Freezer Power Control Relay Read SENS28 MC-FB +5V L Active 29 Read SENS29 AUXAC3-ON-FB +5V L Active 30 Read SENS30 AUXSENS1 31 Read SENS31 AUXSENS2 32 Read SENS32 AUXSENS3 33 Read SENS33 AUXSENS4 34 Read SENS34 AUXSENS5 35 Read SENS35 AUXSENS6 36 Read SENS36 AUXSENS7 37 Read SENS37 AUXSENS8 38 Read SENS38 AUXSENS9 39 Read SENS39 AUXSENS10 40 Read SENS40 AUXSENS11 41 Read SENS41 AUXSENS12 42 Read SENS42 AUXSENS13 43 Read SENS43 AUXSENS14 44 Read SENS44 AUXSENS15 45 Read SENS45 AUXSENS16 46 Read SENS46 AUXSENS17 47 Read SENS47 AUXSENS18 48 Read SENS48 AUXSENS19 49 Read SENS49 AUXSENS20 PCI Bus Interface Connectors (J1/J2) 22 GA4 GA3 GA2 GA2 GA0 GND 21 CLK6 GND RSV RVS RSV GND 20 CLK5 GND RSV GND RSV GND 19 GND GND RXV RSV RSV GND 18 BRSVP2A18 BRSVP2B18 BRSVP2C18 GND BRSVP2E18 GND 17 BRSVP2A17 GND PRST# REQ6# GNT5# GND 16 BRSVP2A16 BRSVP2B16 DEG# GND BRSVPE18 GND 15 BRSVP2A15 GND FAL# REQ5# GNT5# GND 14 AD[35] AD[34] AD[33] GND AD[32] GND 13 AD[38] GND V(I/O) AD[37] AD[36] GND 12 AD[42] AD[41] AD[40] GND AD[39] GND 11 AD[45] GND V(I/O) AD[44] AD[43] GND 10 AD[49] AD[48] AD[47] GND AD[46] GND 9 AD[52] GND V(I/O) AD[51] AD[50] GND 8 AD[56] AD[55] AD[54] GND AD[53] GND 7 AD[59] GND V(I/O) AD[58] AD[57] GND 6 AD[63] AD[62] AD[61] GND AD[60] GND 5 C/BE[5]# GND V(I/O) C/BE[4]# PAR64 GND 4 V(I/O) BRSVP2B4 C/BE[7]# GND C/BE[6]# GND 3 CLK4 GND GNT3# REQ4# GNT4# GND 2 CLK2 CLK3 SYSEN# GNT2# REQ3# GND 1 CLK1 GND REQ1# GNT1# REQ2# GND 25 5V REQ64# ENUM# 3.3V 5V GND 24 AD[1] 5V V(I/O) AS[0] ACK64# GND 23 3.3V AD[4] AD[3] 5V AD[2] GND 22 AD[7] GND 3.3V AD[6] AD[5] GND 21 3.3V AD[9] AD[8] M66EN C/BE[0]# GND 20 AD[12] GND V(I/O) AD[11] AD[10] GND 19 3.3V AD[15] AD[14] GND AD[13] GND 18 SERR# GND 3.3V PAR C/BE[1]# GND 17 3.3V SDONE SBO# GND PERR# GND 16 DEVSEL# GND V(I/O) STOP# LOCK# GND 15 3.3V FRAME# IRDY# GND TRDY# GND 14 13 KEY AREA J1 Connector J2 Connector 12 11 AD[18] AD[17] AD[16] GND C/BE[2]# GND 10 AD[21] GND 3.3V AD[20] AD[19] GND 9 C/BE[3]# IDSEL AD[23] GND AD[22] GND 8 AD[26] GND V(I/O) AD[25] AD[24] GND 7 AD[30] AD[29] AD[28] GND AD[27] GND 6 REQ# GND 3.3V CLK AD[31] GND 5 BRSVP1A5 BRSVP1B5 RST# GND GNT# GND 4 BRSVP1A4 GND V(I/O) INTP INTS GND 3 INTA# INTB# INTC# 5V INTD# GND 2 TCK 5V TMS TDO TDI GND 1 5V -12V TRST# +12V 5V GND B C D E Pin No. A F Motor Driver Signals and Sensors Pin Assignment (J4/J5) 22 LVL1OK H LVL2OK 21 VCC (LVL1OK) 20 LVL1 H AUTO0_1 VCC (LVLV2OK) H LVL2 19 VCC(LVL1) L VCC (AUTO0_1) H AUTO0_2 VCC(LVL2) L VCC (AUTO0_2) TXD1 - TXD2 - GND RXD1 - RXDS - GND RTS1 - RTS2 - GND CTS1 CTS2 GND 18 SM6HOLD L O SM10HOLD L O SENS33 H I DSR1 DSR2 GND 17 SM6TDB L O SM10TDB L O SENS32 H I DTR1 DTR2 GND 16 SM6TDA L O SM10TDA L O SENS31 H I DC30 (TRIG1) L O DC31 (TRIG2) L O GND 15 SM6INB L O SM10INB L O SENS30 H I FB12VGND H I DC13 L O GND 14 SM6INA L O SM10INA L O SENS29 L I SENS38 H I DC12 L O GND 13 SM5HOLD L O SM9HOLD L O SENS28 L I SENS37 H I DC11 L O GND 12 SM5TDB L O SM9TDB L O SENS27 L I SENS36 H I DC10 L O GND 11 SM5TDA L O SM9TDA L O SENS26 L I SENS35 H I DC9 L O GND 10 SM5INB L O SM9INB L O SENS25 L I SENS34 H I DC8 L O GND 9 SM5INA L O SM9INA L O SENS24 L I DC32 L O DC7 L O GND 8 SM4HOLD L O SM8HOLD L O SM13HOLD L O DC29 L O DC6 L O GND 7 SM4TDB L O SM8TDB L O SM13TDB L O DC28 L O DC5 L O GND 6 SM4TDA L O SM8TDA L O SM13TDA L O DC27 L O DC4 L O GND 5 SM4INB L O SM8INB L O SM13INB L O DC26 L O DC3 L O GND 4 SM4INA L O SM8INA L O SM13INA L O DC25 L O DC2 L O GND 3 SM3CCW L O +5V - +5V - DC24 L O DC1 L O GND 2 SM3CW L O +5V - +5V - +5V - SENS49 H I GND 1 SM2HOLD L O +5V - +5V - +5V - SENS48 H I GND 25 SM2TDB L O SM13CCW O SM12HOLD L O DC23 L O SENS47 H I GND 24 SM2TDA L O SM13CW O SM12TDB L O DC22 L O SENS46 H I GND 23 SM2INB L O SM13DLMT I SM12TDA L O DC21 L O SENS45 H I GND 22 SM2INA L O SM7HOLD O SM12INB L O DC20 L O SENS44 H I GND L J5 Connector J5 Connector 21 SM1HOLD L O SM7TDB L O SM12INA L O DC19 L O GND 20 SM1TDB L O SM7TDA L O SM11HOLD L O DC18 L O GND 19 SM1TDA L O SM7INB L O SM11TDB L O DC17 L O GND 18 SM1INB L O SM7INA L O SM11TDA L O DC16 L O AC18K L O GND 17 SM1INA L O SM11F H I SM11INB L O DC15 L O AC1DOWN(PS) L O GND 16 SM6DLMT H I SM11B H I SM11INA L O DC14 L O AC1UP (GH) L O GND 15 SM6PITCH H I SM11PITCH H I SM13PITCH (OK) H I SENS15 H I SENS3 H O GND 14 KEY AREA 13 12 11 SM6HOME H I SM11HOME H I SM13HOME H I SENS14 H I SENS2 H I GND 10 SM5DLMT H I SM10CLASH H I SM12PITCH (OK) H I SENS13 H I SENS1 H I GND 9 SM5PITCH H I SM10PITCH H I SM12HOME H I SENS12 H I SENS43 H I GND 8 SM5HOME H I SM10HOME H I SENS23 H I SENS11 H I SENS42 H I GND 7 SM4DLMT H I SM9PITCH H I SENS22 H I SENS10 H I SENS41 H I GND 6 SM4PITCH H I SM9HOME H I SENS21 H I SENS9 H I SENS40 H I GND 5 SM4HOME H I SM8CLASH H I SENS20 H I SENS8 H I SENS39 H I GND 4 SM3PITCH H I SM8PITCH H I SENS19 H I SENS7 H I AC1OK H I GND 3 SM3HOME H I SM8HOME H I SENS18 H I SENS6 H I AC1S9 H I GND 2 SM2HOME H I SM7PITCH H I SENS17 H I SENS5 H I AC1DOWN (MOVE) H I GND 1 SM1HOME H I SM7HOME H I SENS16 H I SENS4 H I AC1UP (HOME) H I GND A B C D E F Front Panel Interface Connector (CNN1) Connector Pin Map Pin No Unit Name Sensor Signal Value I/O Pin No Unit Name Sensor Signal Value I/O 1 STR1BH SM1HOME L O 51 ISPV-FB SENS15 L O 2 STR2BH SM2HOME L O 52 IR1PV-FB SENS16 L O 3 R2OC SM3HOME L O 53 IR2PV-FB SENS17 L O 4 R2OC SM3PITCH L O 54 ISPBV-FB SENS18 L O 5 SPB SM4HOME L O 55 IR1BPV-FB SENS19 L O 6 SPB SM4PITCH L O 56 IR2BPV-FB SENS20 L O 7 SPB SM4DLMT L O 57 BUZZER-FB SENS21 L O 8 R1BP SM5HOME L O 58 IREFCUP-FB SENS22 L O 9 R1BP SM5PITCH L O 59 AUXDCI-FB SENS23 L O 10 R1BP SM5DLMT L O 60 VP2-FB SENS24 L O 11 R2BP SM6HOME L O 61 VP1-FB SENS25 L O 12 R2BP SM6PITCH L O 62 TWMP-FB SENS26 L O 13 R2BP SM6DLMT L O 63 WPMP-FB SENS27 L O 14 R1ABH SM7HOME L O 64 MC-FB SENS28 L O 15 R1ABH SM7PITCH L O 65 AUXAC3-ON-FB SENS29 L O 16 R1ABZ SM8HOME L O 66 AUXSENS1 SENS30 L O 17 R1ABZ SM8PITCH L O 67 AUXSENS2 SENS31 L O 18 R1ABZ SM8CLASH L O 68 AUXSENS3 SENS32 L O 19 R2ABH SM9HOME L O 69 AUXSENS4 SENS33 L O 20 R2ABH SM9PITCH L O 70 AUXSENS5 SENS34 L O 21 R2ABZ SM10HOME L O 71 AUXSENS6 SENS35 L O 22 R2ABZ SM10PITCH L O 72 AUXSENS7 SENS36 L O 23 R2ABZ SM10CLASH L O 73 AUXSENS8 SENS37 L O 24 NOT USED SM11HOME L O 74 AUXSENS9 SENS38 L O 25 NOT USED SM11PITCH L O 75 AUXSENS10 SENS39 L O 26 NOT USED SM11B L O 76 AUXSENS11 SENS40 L O 27 NOT USED SM11F L O 77 AUXSENS12 SENS41 L O 28 PPBHC SM12HOME L O 78 AUXSENS13 SENS42 L O 29 PPBHC SM12PITCH(OK) L O 79 AUXSENS14 SENS43 L O 30 SP SM13HOME L O 80 AUXSENS15 SENS44 L O 31 SP SM13PITCH(OK) L O 81 AUXSENS16 SENS45 L O 32 SP SM13DLMT L O 82 AUXSENS17 SENS46 L O 33 PWSP AC1UP(HOME) L O 83 AUXSENS18 SENS47 L O 34 PWSP AC1DOWN(MOVE) L O 84 AUXSENS19 SENS48 L O 35 PWSP AC1S9 L O 85 AUXSENS20 SENS49 L O 36 PWSP AC1OK L O 86 37 STR1H-CUV SENS1 L O 87 LVL1OK L O 38 STR2H-CUV SENS2 L O 88 LVL2OK L O 39 STR1BHCUV SENS3 L O 89 GND L O 40 STR2BHCUV SENS4 L O 90 GND L O 41 IAVP-FB SENS5 L O 91 GND L O 42 ESPV-FB SENS6 L O 92 GND L O 43 ERPV-FB SENS7 L O 93 GND L O 44 SWV-FB SENS8 L O 94 O 45 WBSV-FB SENS9 L O 95 O 46 PWV-FB SENS10 L O 96 +5V L O 47 CWV-FB SENS11 L O 97 +5V L O 48 DTG1V-FB SENS12 L O 98 +5V L O 49 DTG2V-FB SENS13 L O 99 +5V L O 50 CDV-FB SENS14 L O 100 +5V L O O Pulse Motor Control Connector CNN2 Pin Map Pin No Signal Name Code Truth Valve I/O 1 SM1 +PO L O 2 SM1 -PO L O 3 SM2 +PO L O 4 SM2 -PO L O 5 SM3 +PO L O 6 SM3 -PO L O 7 SM4 +PO L O 8 SM4 -PO L O 9 SM5 +PO L O 10 SM5 -PO L O 11 SM6 +PO L O 12 SM6 -PO L O 13 SM7 +PO L O 14 SM7 -PO L O 15 SM8 CWP L O 16 CM8 CCWP L O 17 SM9 +PO L O 18 SM9 -PO L O 19 SM10 CWP L O 20 SM10 CCWP L O 21 SM11 +PO L O 22 SM11 -PO L O 23 SM12 +PO L O 24 SM12 -PO L O 25 SM13 +PO L O 26 SM13 -PO L O 27 GND GND O 28 GND GND O 29 GND GND O 30 GND GND O ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Driver Box Assembly Links AC/DC Driver Driver Box Mixer SMD Board ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. AC/DC Driver AC/DC Board AC/DC Driver PCB The AC/DC Driver board drives AC (motors, magnet pumps, etc.) and DC components (magnetic valves, etc.) from the AC/DC or SM_AC/DC controller boards located in the card cage. AC The driver board receives AC motor rotation control and ON/OFF signals from the AC/DC and SM AC/DC Controller boards. The control signals are conveyed by SSR (Solid State Relay) to AC motors, magnetic pumps and magnetic valves. The control signal and the input side of +5 volts are isolated electrically from the output side of the drive circuitry by use of solid state relays mounted in the board. An output circuit is present to provide feedback to the AC/DC and SM AC/DC Controller boards and to check operation of the device (ON/OFF). AC Drives AC Driving Table No Device Code Feedback 1 Wash Unit WSM No 2 Stirrer Arm STRM No 3 Probe Inner Wash Pump PWPM No 4 Cell Wash Pump CWPM No 5 ISE Suction Pump ISPM No 6 ISE Ref Pump IMPM No 7 Detergent Pump DTPM No 8 Cell Wash Vacuum Pump CWP No 9 Cell Dryer Vacuum Pump (spare) CDVP No 10 Outer Wall Wash Pump WPMP No 11 Auto Power Down (Magnet Contact Control) MC1 No 12 Auto Reagent Loading Door Open/Close AUXAC1 No 13 Auto Power Up ON/OFF No DC Drives The DC region is controlled by signals sent from the AC/DC and SM AC/DC Controller boards. The control signal and +5volts is isolated electrically from the output side of the drive circuit by use of a relay mounted on the board. An output circuit provides feedback to the CPU and monitors the relay’s operation. Note Magnetic valves are not equipped with sensors so a feedback signal is prepared in the card cage. DC Driving Table No Device Code Feedback 1 Sampling Pump Valve 1 SPV Y 2 R1 Pump Valve R1PV Y 3 R2 Pump Valve R2PV Y 4 ISE Suction Pinch Valve ISPV Y 5 Acid Valve 1 & Alkaline Valve 1 DTG1V Y 6 Acid Valve 2 & Alkaline Valve 2 DTG2V Y 7 Cell Wash Valve CATCHER Y 8 Sampling Probe Outer Wash Valve ESPV Y 9 Reagent Probe Outer Wash Valve ERPV Y 10 Stirrer Wash Valve SWV Y 11 DIW Supply Valve (Pressurized) WBSV Y 12 DIW Pump (Peristaltic Pump) PWV Y 13 ICT-Level Sensor IREF Y 14 Buzzer BUZZER Y 15 Spare1 CRA CRA N 16 Spare2 AUXDC1 Y 17 Spare3 AUXDC2 Y Input/Output Signals CNN1 Pin Code Description 1 WS UP Cell Wash CW Signal 2 WS DWN Cell Wash CCW Signal 3 WS BK Cell Wash Brake Signal 4 STRZ UP Stirrer Arm CW Signal 5 STRZ DW Stirrer Arm CCW Signal 6 STRZ BK Stirrer Arm Brake Signal 7 PWP UP Probe Inner Wall Wash Pump CW Signal Value L 8 PWP DWN Probe Inner Wall Wash Pump CCW Signal 9 PWP BK Probe Inner Wall Wash Pump Brake Signal 10 AUXAC1 11 AUXAC1 12 AUXAC1 13 NC 14 WCP ON Cell Wash Pump CW Signal 15 WCP BK Cell Wash Pump Brake Signal 16 ICT ON ISE Suction Pump CW Signal 17 ICT BK ISE Suction Pump Brake Signal 18 ICTR ON ISE Ref Pump CW Signal 19 ICTR BK ISE Ref Pump Brake Signal 20 DTGP ON Detergent Pump CW Signal 21 DTGP BK Detergent Pump Brake Signal 22 NC 23 VP2 Cell Vacuum Pump ON Signal (Vacuum) 24 VP1 Spare ON Signal (Cell Dryer Pump) 25 WPMP Outer Wall Pump ON Signal 26 MC1 Auto Power OFF-ON Signal 27 NC 28 ISPV Sample Pump Valve 1&2 ON Signal 29 IR1PV R1 Pump Valve ON signal 30 IR2PV R2 Pump Valve ON Signal 31 IAV ISE Suction Pinch Valve ON Signal 32 DTG1V Acid & Alkaline Detergent Valve 1 ON Signal 33 DTG2V Acid & Alkaline Detergent Valve 2 ON Signal 34 CWV Cell Wash Valve 35 ESPV Sample Probe Valve ON Signal 36 ERPV R-Probe Outer Wash Valve ON Signal 37 SWV Stirrer Wash Valve ON Signal 38 WBSV DIW Supply Valve ON Signal 39 PWV DIW Supply Pump (Peristaltic Pump) ON 40 IREF ICT-Level Sensor ON Signal 41 BUZZER Buzzer ON Signal 42 AUXDC1 Spare 43 AUXDC2 Spare 44 AUTO1 CRA Door Open/Close 1 Signal 45 AUTO2 CRA Door Open/Close 2 Signal 46 NC 47 NC 48 NC 49 NC 50 NC 51 NC 52 ISPV FB Sample Pump Valve 1 & 3 FB Signal 53 IR1PV FB R1 Pump Valve FB Signal 54 IR2PV FB R2 Pump Valve FB Signal 55 IAV FB ISE Suction Pinch Valve FB Signal 56 DTG1V FB Acid & Alkaline Valve 1 FB Signal 57 DTG2V FB Acid & Alkaline Valve 2 FB Signal 58 CWV FB Cell Wash Valve FB Signal 59 ESPV FB Sample Probe Outer Wash Valve FB Signal 60 ERPV FB R-Probe Outer Wall Wash Valve FB Signal 61 SWV FB Stirrer Wash Valve FB Signal 62 WBSV FB DIW Supply Valve FB signal 63 PWV FB DIW Supply Pump (Peristaltic Pump) FB Signal 64 IREF FB ICT-Level Sensor FB Signal 65 BUZZER Buzzer FB Signal 66 AUXDC1 Spare 1 FB Signal 67 AUXDC2 Spare 2 FB Signal 68 NC 69 GND (+12V) Feedback GND (+12V) 70 GND (+12V) Feedback GND (+12V) 71 GND GND 72 GND GND 73 GND GND 74 GND GND 75 GND GND 76 +5V +5V Power Supply 77 +5V +5V Power Supply 78 +5V +5V Power Supply 79 +5V +5V Power Supply 80 +5V +5V Power Supply AC Drive Signal Output CNN2 – CNN7, CNN14 CNN 2 1 Cell Wash Motor 2 3 4 Stirrer Arm Motor 5 6 CNN 3 1 Probe Inner Wall Wash Pump Motor 2 3 4 Cell Wash Pump Motor 5 6 CNN 4 1 ISE Suction Pump Motor 2 3 4 ISE Ref Pump Motor 5 6 CNN 5 1 Detergent Pump Motor 2 3 4 Reagent Supply Door Open/Close Motor 5 6 CNN 6 1 Cell Wash Vacuum Pump 2 CNN 7 1 Cell Wash Dryer Pump 2 3 Outer Wall Wash Pump 4 5 Auto Power Down (Magnet Contact) 6 CNN 14 1 2 3 Auto Power On/Off DC Drive Signal Output CNN 8 – CNN 10 CNN 8 1 Sampling Pump Valve 1&2 2 3 R1 Pump Valve 4 5 R2 Pump Valve 6 7 ISE Suction Pinch Valve 8 9 Acid Valve 1 & Alkaline Detergent Valve 1 10 11 Acid Valve 2 & Alkaline Detergent Valve 2 12 CNN 9 1 Cell Wash Valve 2 3 Sampling Probe Outer Wall Wash Valve 4 5 Reagent Probe Outer Wall Wash Valve 6 7 Stirrer Wash Valve 8 9 DIW Supply Valve 10 11 DIW Supply Pump 12 CNN 10 1 ICT-Level Sensor 2 3 Buzzer 4 5 Spare 1 (CRA) 6 7 Spare 2 8 9 10 Spare 3 Power Input CNN 11, CNN 12, CNN 15 CNN 11 1 AC100V 2 3 NC CNN 12 1 DC+12V 2 CNN 15 1 AC100V 2 3 NC Automatic Power ON/OFF Control Signal CNN 13 1 Automatic Power ON/OFF Control Signal 2 ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064. All rights reserved. Driver Box The Driver Box Assembly contains driver boards for motors and magnetic valves. The following boards are contained in the Driver Box. • SMD Board–The SMD Board receives stepper motor control signals from the SMC and SM AC/DC boards. Power for each driver is supplied from the power supply. Drive signals for each stepping motor is provided by each mounted driver circuit. • AC/DC Board–The AC/DC Driver Board receives ON/OFF and motor rotation control signals from the AC/DC and SM AC/DC Controller boards. Drive signals for 100VAC motors, 12VDC magnetic pumps and magnetic valves, are converted by SSR; with the relay mounted on the board. • STR-Driver 2 Board–The STR-Driver (Mixer) 2 Board receives control signals for the Piezo Stirrers from the SMC and SM AC/DC Controller boards, and provide output signals for driving the Piezo Stirrer for each STR unit. The AC/DC Controller board provides control signals for vertical (up/down) movement of the Mixer Assembly. SMD Connector Position SMD Connector Function CN1 I/F Signals CN2 Reaction Carousel Rotate Motor CN3 Sample Arm U/D Motor CN4 Reagent Inner Carousel Motor CN5 Reagent Outer Carousel Motor CN6 Sample Arm Rotate Motor CN7 R1 Arm U/D Rotate Motor CN8 R2 Arm U/D Rotate Motor CN9 Stirrer 1 and 2 Rotate Motor CN10 ISE Arm U/D and Rotate Motor CN11 Sample Pump and R1 Pump Motors CN12 R2 and HCW Pump Motors CN13 CRA CN14 CRA CN15 CN16 CN17 CN18 CN19 CN20 CN21 CN22 CN23 CN24 AC/DC Connector Position AC/DC Driver Connector Function CN1 I/F Signals CN2 Cell Wash Motor/Stirrer Arm Motor CN3 Probe Inner Wash Pump Motor/Cell Wash Pump Motor CN4 ISE Suction Pump Motor CN5 Detergent Pump Motor/Reagent Supply Door Open/Close Motor CN6 Cell Wash Vacuum Pump CN7 Cell Wash Dryer Pump/Outer Wall Wash Pump/Auto Power Down CN8 S, R1, R2 Pump Valves/ISE Suction Pinch Valve/Acid Alk Det Valves 1 and 2 CN9 Wash Valves/DIW Supply Valve and Pump CN10 ICT Level Sensor/Buzzer/CRA CN11 AC100V CN12 DC+12V CN13 Auto Power On/Off CN14 AC100V CN15 On/Off STR-Driver2 STR-Driver2 Connector Function CN1 STR-Driver2 Control Signal I/F CN2 Piezo Stirrer Drive Signal Output CN3 +24VAC CN4 ±15VAC Input/Output Connectors Driver Box Assembly External Connector Position Connector Function SM1 SDM Control Signal I/F SM2,3 5-phase Stepping Motor Drive Output SM4,5 5-phase Stepping Motor Drive Output SM6 5-phase Stepping Motor Drive Output SM7,8 2-phase Stepping Motor Drive Output SM8,9 2-phase Stepping Motor Drive Output SM10,11 2-phase Stepping Motor Drive Output SM11,12 2-phase Stepping Motor Drive Output SM13,14 2-phase Stepping Motor Drive Output SM15-17 DC+24V Input SM18-21 DC+24V Input SM22 Solenoid DC+24V Output DR_FAN AC100V for Cooling Fan AD1 AC/DC Control Signal I/F AD2,3 AC Drive Signal Output AD4,5 AC Drive Signal Output AD6,7 AC Drive Signal Output AD8 AC Drive Signal Output AD9 AC Drive Signal Output AD10 AC Drive Signal Output AD11 AC100V Input AD12 DC+12V Input STR1 STR-DRIVER 2 Control Signal I/F STR2 Piezo Stirrer Drive Signal Output STR3,4 AC+24V/DC+15V Input FB Not used ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Mixer Mixer Block Diagram Mixer Driver Board Functional Diagram Link STR-Driver2 (FSTR2A) 7A_2044b NEW STR-Driver (FSTR2C) 7A_2078a Functional Diagram Mixer Block Diagram Link 7A_2077a Test Points TP1 +15V LED 1 (voltage supplied) TP2 for analog IC's GND TP3 -15V LED 2 (voltage supplied) TP4 +50V Not monitored by LED TP5 for driving the Piezo stirrers GND TP6 -50V Not monitored by LED TP7 +5V LED 5 (voltage generated on board) TP8 GND TP9 +3.3V LED 6 (voltage generated on board) TP10 for the FPGA GND TP11 <1OUT> TP12 <2OUT> Switch Settings SW1 1 ON 2 ON 3 OFF 4 OFF SW2 1 ON 2 ON 3 OFF 4 OFF Address Settings Address SW3 1 SW4 4 SW5 5 SW5 5 SW7 1 SW8 4 SW9 5 SW10 5 Cable Connection Panel Cable Label Connection Label Control Function STR-A ST1AM Stirrer 1 rotate motor STR-B ST2AM Stirrer 2 rotate motor STR-D STRD Stirrer vibration drive STR-E STRM Stirrer Mixer Elevator Motor SE-S STRE-S Stirrer Elevator Motor Sensors STR-S STR-S Stirrer Rotation Home and Cuvette Sensors ARCHITECT c4000 Service and Support Manual (Version 204733-108) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. SMD Board The SMD (Stepping Motor Driver) Board receives control signals from the SMC and/or SM_AC/DC Control Boards and 24VDC from the power supply to provide drive signals for each stepper motor. SMD Block Diagram SMD Board I/F Signals CNN 1 Pin No Signal Name Active Note 1 RCCW L Reaction Carousel Rotation Motor CW 2 RCCCW L Reaction Carousel Rotation Motor CCW 3 RICCW L Reagent Inner Carousel Motor CW 4 RICCCW L Reagent Inner Carousel Motor CCW 5 ROCCW L Reagent Outer Carousel Motor CW 6 ROCCCW L Reagent Outer Carousel Motor CCW 7 SAZCW L Sample Arm Up/Down Motor CW 8 SAZCCW L Sample Arm Up/Down Motor CCW 9 SAHCW L Sample Arm Rotate Motor CW 10 SAHCCW L Sample Arm Rotate Motor CCW 11 R1ZNA L R1 Arm Up/Down Motor NA 12 R1ZNB L R1 Arm Up/Down Motor NB 13 R1ZTDA L R1 Arm Up/Down Motor TDA 14 R1ZTDB L R1 Arm Up/Down Motor TDB 15 R1ZHOLD L R1 Arm Up/Down Motor HOLD 16 R1HNA L R1 Arm Rotate Motor NA 17 R1HNB L R1 Arm Rotate Motor NB 18 R1HTDA L R1 Arm Rotate Motor TDA 19 R1HTDB L R1 Arm Rotate Motor TDB 20 R1HHOLD L R1 Arm Rotate Motor HOLD 21 R2ZNA L R2 Arm Up/Down Motor NA 22 R2ZNB L R2 Arm Up/Down Motor NB 23 R2ZTDA L R2 Arm Up/Down Motor TDA 24 R2ZTDB L R2 Arm Up/Down Motor TDB 25 R2ZHOLD L R2 Arm Up/Down Motor HOLD 26 R2HNA L R2 Arm Rotate Motor NA 27 R2HNB L R2 Arm Rotate Motor NB 28 R2HTDA L R2 Arm Rotate Motor TDA 29 R2HTDB L R2 Arm Rotate Motor TDB 30 R2HHOLD L R2 Arm Rotate Motor HOLD 31 STR1HNA L Stirrer 1 Rotate Motor NA 32 STR1H/NB L Stirrer 1 Rotate Motor NB 33 STR1HTDA L Stirrer 1 Rotate Motor TDA 34 STR1HTDB L Stirrer 1 Rotate Motor TDB 35 STR1HHOLD L Stirrer 1 Rotate Motor HOLD 36 STR2HINA L Stirrer 2 Rotate Motor NA 37 STR2HINB L Stirrer 2 Rotate Motor NB 38 STR2HTDA L Stirrer 2 Rotate Motor TDA 39 STR2HTDB L Stirrer 2 Rotate Motor TDB 40 STR2HHOLD L Stirrer 2 Rotate Motor HOLD 41 IAZINA L ISE Arm Up/Down Motor NA 42 IAZINB L ISE Arm Up/Down Motor NB 43 IAZTDA L ISE Arm Up/Down Motor TDA 44 IAZ TDB L ISE Arm Up/Down Motor TDB 45 IAZ HOLD L ISE Arm Up/Down Motor HOLD 46 IAHINA L ISE Arm Rotate Motor NA 47 IAHINB L ISE Arm Rotate Motor NB 48 IAHTDA L ISE Arm Rotate Motor TDA 49 IAH TDB L ISE Arm Rotate Motor TDB 50 IAH HOLD L ISE Arm Rotate Motor HOLD 51 SPINA L Sample Pump Motor NA 52 SPINB L Sample Pump Motor NB 53 SPTDA L Sample Pump Motor TDA 54 SPTDB L Sample Pump Motor TDB 55 SPHOLD L Sample Pump Motor HOLD 56 R1PINA L R1 Pump Motor NA 57 R1PINB L R1 Pump Motor NB 58 R1P TDA L R1 Pump Motor TDA 59 R1PTDB L R1 Pump Motor TDB 60 R1PHOLD L R1 Pump Motor HOLD 61 R2PINA L R2 Pump Motor NA 62 R2PINB L R2 Pump Motor NB 63 R2P TDA L R2 Pump Motor TDA 64 R2PTDB L R2 Pump Motor TDB 65 R2PHOLD L R2 Pump Motor HOLD 66 PPHCNA L HCW Pump Motor NA 67 PPHCNB L HCW Pump Motor NB 68 PPHCTDA L HCW Pump Motor TDA 69 PPHCTDB L HCW Pump Motor TDB 70 PPHCHOLD L HCW Pump Motor HOLD 71 AUTO1NA L Reagent Handler (1) Motor NA 72 AUTO1NB L Reagent Handler (1) Motor NB 73 AUTO1TDA L Reagent Handler (1) Motor TDA 74 AUTO1TDB L Reagent Handler (1) Motor TDB 75 AUTO1HOLD L Reagent Handler (1) Motor HOLD 76 AUTO2NA L Reagent Handler (2) Motor NA 77 AUTO2NB L Reagent Handler (2) Motor NB 78 AUTO2TDA L Reagent Handler (2) Motor TDA 79 AUTO2TDB L Reagent Handler (2) Motor TDB 80 AUTO2HOLD L Reagent Handler (2) Motor HOLD 81 AUTO3NA L Reagent Handler (3) Motor NA 82 AUTO3NB L Reagent Handler (3) Motor NB 83 AUTO3TDA L Reagent Handler (3) Motor TDA 84 AUTO3TDB L Reagent Handler (3) Motor TDB 85 AUTO3HOLD L Reagent Handler (3) Motor HOLD 86 AUXNA L Spare Motor NA 87 AUXNB L Spare Motor NB 88 AUXTDA L Spare Motor TDA 89 AUXTDB L Spare Motor TDB 90 AUXHOLD L Spare Motor HOLD 91 GND L GND (+5V) 92 GND L GND (+5V) 93 GND L GND (+5V) 94 GND L GND (+5V) 95 GND L GND (+5V) 96 GND L GND (+5V) 97 +5V L +5V Power 98 +5V L +5V Power 99 +5V L +5V Power 100 +5V L +5V Power 5-Phase Stepping Motor Drive Signals No Unit Name Code 1 Reaction Carousel Motor RUM 2 Sample Arm Up/Down Motor SAEM 3 Reagent Inner Rotate Motor RICM 4 Reagent Outer Rotate Motor ROCM 5 Sample Arm Theta Motor SAAM CNN 2 Pin 1 Note Reaction Carousel Rotate Motor 2 3 4 5 CNN 3 Pin 1 Note Sample Arm Up/Down Motor 2 3 4 5 CNN 4 Pin 1 Note Reagent Inner Carousel Motor 2 3 4 5 CNN 5 Pin 1 Note Reagent Outer Carousel Motor 2 3 4 5 CNN 6 Pin Note 1 Sample Arm Rotate Motor 2 3 4 5 2-Phase Stepping Motor Drive Table No Unit Name Code 1 R1 Arm Up/Down Motor R1AEM 2 R1 Arm Rotate Motor R1AAM 3 R2 Arm Up/Down Motor R2AEM 4 R2 Arm Rotate Motor R2AAM 5 R1 Stirrer Rotate Motor ST1AM 6 R2 Stirrer Rotate Motor ST1AM 7 ISE Arm Up/Down Motor IAEM 8 ISE Arm Rotate Motor IAAM 9 Sampling Pump Motor SPM 10 R1 Pump Motor R1PM 11 R2 Pump Motor R2PM 12 HCW Pump Motor HCWPM 13 Reagent Loading Motor (1) 14 Reagent Loading Motor (2) 15 Reagent Loading Motor (3) 16 Spare Reagent Loading Motor (4) 2-Phase Stepping Motor Drive Signals CNN 7 Pin 1 Note R1 Arm Up/Down Motor 2 3 4 5 6 7 8 9 10 11 12 CNN8 R1 Arm Rotate Motor Pin 1 Note R2 Arm Up/Down Motor 2 3 4 5 6 7 R2 Arm Rotate Motor 8 9 10 11 12 CNN9 Pin 1 Note Stirrer 1 Rotate Motor 2 3 4 5 6 7 Stirrer 2 Rotate Motor 8 9 10 11 12 CNN10 Pin 1 Note ISE Arm Up/Down Motor 2 3 4 5 6 7 8 9 10 11 ISE Arm Rotate Motor 12 CNN11 Pin 1 Note Sampling Pump Motor 2 3 4 5 6 7 R1 Pump Motor 8 9 10 11 12 CNN12 Pin 1 Note R2 Pump Motor 2 3 4 5 6 7 HCW Pump Motor 8 9 10 11 12 CNN13 Pin 1 Note Reagent Auto Loading Motor (1) 2 3 4 5 6 7 Reagent Auto Loading Motor (2) 8 9 10 11 12 CNN14 Pin 1 Note Reagent Auto Loading Motor (3) 2 3 4 5 6 7 Reagent Auto Loading Motor (4) 8 9 10 11 12 Power Inputs CNN15–CNN21 CNN15 Pin Note CNN16 Pin Note CNN17 Pin Note CNN18 Pin Note CNN19 Pin Note CNN20 Pin Note CNN21 Pin Note Solenoid +24V Outputs CNN 22 Pin Note ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Miscellaneous Boards Links CNN Board ICT Pre Amp Board IREF Level Detect Board LED Boards LLS Board Thermal Bath Controller (TBC) Board ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. CNN Board The CNN Board is the distribution board for branch wiring between the card cage and other boards or devices. The board maintains +5, +12 and ±15VDC for supply to sensors, pressure sensors and power for level sensor board. The TLLS board and +15VDC-DC converter for the TLLS-Pre Board are mounted on the CNN Board. CNN Board Callouts CNN1 (from power supply via Card Cage) Pin Signal Name I/O Note 1 +5V I +5V 2 GND O Pin Signal Name I/O Note 36 +12V I ±12V CNN5 (from power supply via Card Cage) 37 I 38 I 39 I 40 I 41 -12V I 42 I 43 I 44 I 45 I 46 12VGND I 47 I 48 I 49 I 50 I CNN6 (from power supply via Card Cage) Pin Signal Name I/O Note 36 +12V I +/−12V 37 I 38 I 39 I 40 I 41 -12V I 42 I 43 I 44 I 45 I 46 12VGND I 47 I 48 I 49 I 50 I CNN11>CNN3 (Pumps) Pin Signal Signal Name 1 +5v +5v 2 GND GND 3 Pitch SPMIR 4 +5v +5v 5 GND GND 6 Home SPHOME 7 +5v +5v 8 GND GND 9 Down Limit SPDLMT 10 +5v +5v 11 GND GND 12 Pitch R1PMIR 13 +5v +5v 14 GND GND Active I/O Note Pin Signal Active I/O Sample Pump H I 7 SPMIR H I H I 6 SPHOME H I H I 8 SPDLMT H I 10 R1PMIR H I R1 Pump H I 15 Home R1PHOME 16 +5v +5v 17 GND GND 18 Down Limit R1PDLMT 19 +5v +5v 20 GND GND 21 Pitch R2PMIR 22 +5v +5v 23 GND GND 24 Home R2PHOME 25 +5v +5v 26 GND GND 27 Down Limit R2PDLMT 28 NC 29 NC 30 NC 31 NC 32 NC 33 NC 34 NC H I 9 R1PHOME H I H I 11 R1PDLMT H I R2 Pump H I 13 R2PMIR H I H I 12 R2PHOME H I H I 14 R2PDLMT H I Active I/O CNN12>CNN3 (Probes and Level Sense) Pin Signal Signal Name 1 +15V +15V AGND AGND -15V -15V Active I/O Note Pin Signal 2 3 4 5 6 7 VCCSAAAUTOZERO +5V 53 VCCSAAAUTOZERO 8 SAAUTOZERO S Probe LLS Auto Zero 46 SAAUTOZERO 9 VCCSALVL S Probe LLS (+) Differential 49 VCCSALVL 10 SALVL S Probe LLS (-) Differential 50 SALVL 11 VCCSALVLOK S Probe LLS (+) Integral 51 VCCSALVLOK 12 SALVLOK S Probe LLS (-) Integral 52 SALVLOK 13 VCCR1AAUTOZERO +5V 59 VCCR1AAUTOZERO 14 R1AAUTOZERO R1A Probe LLS Auto Zero 17 R1AAUTOZERO 15 VCCR1ALVL R1A Probe LLS (+) Differential 55 VCCR1ALVL 16 R1ALVL R1A Probe LLS (-) 56 R1ALVL O L O I H I I H I O L O I H I Differential 17 VCCR1ALVLOK R1A Probe LLS (+) Integral 57 VCCR1ALVLOK I 18 R1ALVLOK R1A Probe LLS (-) Integral 58 R1ALVLOK 19 VCCR2AAUTOZERO +5V 65 VCCR2AAUTOZERO 20 R2AAUTOZERO R2A Probe LLS Auto Zero 22 R2AAUTOZERO 21 VCCR2ALVL R2A Probe LLS (+) Differential 61 VCCR2ALVL 22 R2ALVL R2A Probe LLS (-) Differential 62 R2ALVL 23 VCCR2ALVLOK R2A Probe LLS (+) Integral 63 VCCR2ALVLOK 24 R2ALVLOK R2A Probe LLS (-) Integral 64 R2ALVLOK 25 SALVLGUV+5V 68 SALVLGUV+5V 26 SALVLGU 67 SALVLGU L Pin Signal Active I/O H I O L O I H I I H I CNN13>CNN2 (Degasser and TBC) Pin Signal Signal Name Active I/O Note 1 +5V +5V 2 +5V +5V 3 BathWater Htr ON/OFF IHTRON L O 77 IHTRON L O 4 WaterBath CyclePump ON/OFF IPMON L O 76 IPMON L O 5 NC 6 Water Bath Fan CntrlON/OFF IFANON L O 80 IFANON L O 7 Water Bath Mtr Vlv Open IMVOPEN L O 78 IMVOPEN L O 8 Water Bath Mtr Vlv Clsd IMVSHUT L O 79 IMVSHUT L O 9 Wtr Bth Fan Ctrl Thrm IVRSTR+ I 89 IVRSTR+ I 10 Wtr Bth Fan Ctrl Thrm IVRSTR- I 90 IVRSTR- I 11 +5V +5V 12 GND GND 13 Disp PWPUP 7 PWPUP H I 14 +5V +5V 15 GND GND 16 Aspir PWPDOWN 8 PWPDOWN H I 17 Degasser Fail Detection DALARMV I 69 DALARMV I 18 Degasser DALARM I 68 DALARM I TBC Probe H Wash Pump H Degasser Fail Detection 19 NC 20 NC CNN14>CNN2 Pin Signal Signal Name 1 +5V +5V 2 GND GND 3 Disp DPUP 4 +5V +5V 5 GND GND 6 Aspir DPDOWN 7 +5V +5V 8 GND GND 9 Blank ESMCC 10 +5V +5V 11 GND GND 12 Blank ESB3 13 +5V +5V 14 GND GND 15 Blank ESB14 16 +5V +5V 17 GND GND 18 Disp CWPUP 19 +5V +5V 20 GND GND 21 Aspir CWPDOWN 22 GND GND 23 Full Water FFS2UPPERLIM 24 GND GND 25 Low Level FFS4OPENL 26 GND GND 27 Full Water FFS3OPENH 28 GND GND 29 Full Water FFS1LW 30 GND GND 31 DTFHW 32 GND GND 33 Full Water FFS5AUXFSW 34 NC CNN15 >CNN3 (Sample Arm) Active I/O Note Pin Signal Active I/O Detergent Pump H I 12 DPUP H I H I 13 DPDOWN H I 70 ESMCC H I 72 ESB3 H I 71 ESB14 H I ISE Ref Solution H I Acid Detergent H I Alkaline Detergent H I Cell Wash Pump H I 14 CWPUP H I H I 15 CWPDOWN H I 63 FFS2UPPERLIM I 65 FFS4OPENL I 64 FFS3OPENH I 62 FFS1LW I 67 DTFHW I 66 FFS5AUXFSW I I I I Open Tank Upper Limit Open Tank Low Level Open Tank Full Level LCW Full Detection I HCW Full Detection I Spare Float SW I Pin Signal Signal Name 1 +5V +5V 2 GND GND 3 Rotate Pitch SAAM1R 4 +5V +5V 5 GND GND 6 Rotate Home SAAHOME 7 +5V +5V 8 GND GND 9 U/D Pitch SAEM1R 10 +5V +5V 11 GND GND 12 U/D Home SAEHOME 13 +5V +5V 14 GND GND 15 +5V +5V 16 Crash SACLASH Active I/O Note Pin Signal Active I/O Sample Arm H I 45 SAAM1R H I H I 44 SAAHOME H I H I 47 SAEM1R H I H I 46 SAEHOME H I H I 48 SACLASH H I CNN16 >CNN2, CNN3 and CNN4 (Reagent Supply) Pin Signal Signal Name 1 +5V +5V 2 GND GND 3 Inner Pitch R1CIPITCH 4 +5V +5V 5 GND GND 6 Inner Home R1CIHOME 7 +5V +5V 8 GND GND 9 Outer Pitch R1COPITCH 10 +5V +5V 11 GND GND 12 Outer Home R1COHOME 13 Temp Sensor R1CTEMP- 14 Temp Sensor R1CTEMP+ 15 +5V +5V 16 GND GND 17 Cover Present Sensor RCVROK 18 +5V +5V 19 GND GND Active I/O Note Cable Reagent Supply CNN3 Pin Signal Active I/O H I 36 R1CIPITCH H I H I 35 R1CIHOME H I H I 39 R1COPITCH H I H I 38 R1COHOME H I 86 R1CTEMP- 85 R1CTEMP+ 59 RCVROK H I CNN2 H I CNN4 20 Outer Rotate SW 21 +5V +5V 22 GND GND 23 Inner Rotate SW 24 +12V +12V 25 +12VGND +12VGND 26 Bottle Sensor BTTLSNS 27 +5V +5V 28 GND GND 29 Cup Sensor CUPSNSRO 30 NC 31 NC 32 NC 33 NC 34 NC 35 +5V +5V 36 GND GND 37 Cover Opn/Cls Sensor Open COVER OPEN 38 +5V +5V 39 GND GND 40 Cover Opn/Cls Sensor Close COVER CLOSE CNN2 95 BTTLSNS I H I 96 CUPSNSRO H I H I 10 COVEROPEN H I H I H I 9 COVER CLOSE CNN17>CNN3 (R1 Arm) Pin Signal Signal Name 1 +5V +5V 2 GND GND 3 Rotat Pitch R1AM1R 4 +5V +5V 5 GND GND 6 Rotat Home R1AHOME 7 +5V +5V 8 GND GND 9 U/D Pitch R1EM1R 10 +5V +5V 11 GND GND 12 U/D Home R1EHOME 13 +5V +5V Active I/O Note Pin Signal Active I/O R1 Arm H I 16 R1AM1R H I H I 15 R1AHOME H I H I 18 R1EM1R H I H I 17 R1EHOME H I 14 GND GND 15 +5V +5V 16 Crash R1CLASH H I Active I/O 19 R1CLASH H I CNN18>CNN2, CNN3 and CNN4 Pin Signal Signal Name 1 +5V +5V 2 GND GND 3 Up STEUP 4 +5V +5V 5 GND GND 6 Down STEDOWN 7 +5V +5V 8 GND GND 9 Down OK STEDOK 10 +5V +5V 11 GND GND 12 STR1 Rotat Home STA1HOME 13 +5V +5V 14 GND GND 15 STR1 Rotat Cuv Sense SENS1 16 +5V +5V 17 GND GND 18 STR 2 Rotat Home STA2HOME 19 +5V +5V 20 GND GND 21 STR2 Rotat Cuv Sense SENS2 22 +5V +5V 23 GND GND 24 Up WSUP 25 +5V +5V 26 GND GND 27 Down WSDOWN 28 +5V +5V 29 GND GND 30 Down OK WSDOWNOK 31 +5V +5V 32 GND GND 33 Pitch RUM1R Note Cable Stirrer CNN2 Pin Signal Active I/O H I 4 STEUP H I H I 5 STEDOWN H I H I 6 STEDOK H I 4 STA1HOME H I 37 SENS1 H I 5 STA2HOME H I 38 SENS2 H I CNN3 H I CNN4 H I CNN3 H I CNN4 H I Wash Elevator CNN2 H I 1 WSUP H I H I 2 WSDOWN H I H I 3 WSDOWNOK H I 34 RUM1R H I Reaction Cuvette H I CNN3 34 +5V +5V 35 GND GND 36 Home RUHOME 37 GND GND 38 Water Bath Full Detect 39 40 H I 33 RUHOME H I BATHLVL I 58 BATHLVL I Water Bath Temp Sensor BATHTMP- I 84 BATHTMP- I Water Bath Temp Sensor BATHTMP+ I 83 BATHTMP+ I CNN19>CNN3 (R2 Arm) Pin Signal Signal Name 1 +5V +5V 2 GND GND 3 Rotate Pitch R2AM1R 4 +5V +5V 5 GND GND 6 Rotate Home R2AHOME 7 +5V +5V 8 GND GND 9 U/D Pitch R2EM1R 10 +5V +5V 11 GND GND 12 U/D Home R2EHOME 13 +5V +5V 14 GND GND 15 +5V +5V 16 Crash R2CLASH Active I/O Note Pin Signal Active I/O R2 Arm H I 21 R2AM1R H I H I 20 R2AHOME H I H I 23 R2EM1R H I H I 22 R2EHOME H I H I 24 R2CLASH H I Active I/O Note Pin Signal Active I/O Reagent Barcode Reader 11 TXD2 13 RTS2 L O CNN20 >CNN7 (Reagent Barcode Reader) Pin Signal Signal Name 1 TXD TXD2 2 RTS RTS2 3 NC NC 4 RXD RXD2 12 RXD2 5 CTS CTS2 14 CTS2 6 NC 7 GND GND 8 TRIGGER TRIGGER2 15 TRIGGER2 L O 9 +5V +5V 10 +5V +5V 11 GND GND 12 NC 13 NC 14 NC 15 NC 16 NC CNN21>CNN7 (LUI Serial) Pin Signal Signal Name 1 TXD TXD4 2 RTS 3 NC 4 Active I/O Note Pin Signal LUI Serial 21 TXD4 RTS4 23 RTS4 RXD RXD4 22 RXD4 5 CTS CTS4 24 CTS4 6 NC 7 GND GND 8 TRIGGER TRIGGER4 25 9 +5V +5V 10 +5V +5V 11 GND GND 12 NC 13 NC 14 NC 15 NC 16 NC Note CLI Serial L O Active I/O Active I/O TRIGGER4 L O Pin Signal Active I/O 26 CD5 CNN22>CNN7 (CLI Serial) Pin Signal Signal Name 1 Carrier Detect CD5 2 Receive Data RXD5 27 RXD5 3 Transmit Data TXD5 28 TXD5 4 Data Terminal Ready DTR5 29 DTR5 5 GND GND 6 Data Set Ready DSR5 30 DSR5 7 Request to Send RTS5 31 RTS5 8 Clear to CTS5 32 CTS5 GND Send 9 Ring Indicator 10 NC CI5 33 CI5 CNN23>CNN2 (Status LEDs) Pin Signal Signal Name Active I/O Note Pin Signal Active I/O 1 CPULED0 CPULED0 L O Status LED 40 CPULED0 L O 2 CPULED1 CPULED1 L O 41 CPULED1 L O 3 CPULED2 CPULED2 L O 42 CPULED2 L O 4 CPULED3 CPULED3 L O 43 CPULED3 L O 5 CPULED4 CPULED4 L O 44 CPULED4 L O 6 CPULED5 CPULED5 L O 45 CPULED5 L O 7 CPULED6 CPULED6 L O 46 CPULED6 L O 8 CPULED7 CPULED7 L O 47 CPULED7 L O 9 CPULED8 CPULED8 L O 48 CPULED8 L O 10 CPULED9 CPULED9 L O 49 CPULED9 L O 11 CPULED10 CPULED10 L O 50 CPULED10 L O 12 CPULED11 CPULED11 L O 51 CPULED11 L O 13 CPULED12 CPULED12 L O 52 CPULED12 L O 14 CPULED13 CPULED13 L O 53 CPULED13 L O 15 CPULED14 CPULED14 L O 54 CPULED14 L O 16 CPULED15 CPULED15 L O 55 CPULED15 L O 17 +5V +5V 18 +5V +5V 19 GND GND 20 GND GND 21 GND GND 22 GND GND 23 NC 24 NC 25 NC 26 NC Active I/O Note Pin Signal Active I/O Shutter 36 SHUTTER ON L O CNN24>CNN7 (Optics) Pin Signal Signal Name 1 +5V 2 Optics Shutter SHUTTER ON 3 LED VCC LEDVCC 4 GND 5 +5V 6 Optics Trigger OPT-TRIGGE Optics Trigger 37 LEDVCC 39 OPT-TRIGGE O H O Signal 7 GND CNN25>CNN2 and CNN6 Pin Signal 1 +12V 2 +/12VGND 3 -12V 4 NC 5 S Pressure Sensor Sig Pos 6 S Pressure Sensor Sig Neg 7 +12V 8 +/12VGND 9 -12V 10 NC 11 Signal Name Active I/O Note Cable Pin Signal Pressure Sensor CNN6 SASEN1 POS 1 SASEN1 POS SASEN1 NEG 2 SASEN1 NEG R1 Pressure Sensor Sig Pos R1ASEN POS 3 R1ASEN POS 12 R1 Pressure Sensor Sig Neg R1ASEN NEG 4 R1ASEN NEG 13 +12V 14 +/12VGND 15 -12V 16 NC 17 R2 Pressure Sensor Sig Pos R2ASEN POS 5 R2ASEN POS 18 R2 Pressure Sensor Sig Neg R2ASEN NEG 6 R2ASEN NEG 19 TLLS SLLS 33 SLLS 20 TLLSG SLLSG 34 SLLSG 21 +5V 22 GND 23 ISE Suct Pump UP ISPUP 16 ISPUP 24 +5V 25 GND 26 ISE Suct ISPDOWN 17 ISPDOWN Sample TLLS ISE Pump CNN2 Active I/O Pump DOWN 27 +5V 28 GND 29 ISE Ref Pump UP 30 NC 31 NC 32 NC 33 NC 34 NC IMPUP 18 IMPUP CNN26>CNN2 (STR Driver) Pin Signal 1 +5V 2 +5V 3 Stirrer 1 On 4 Signal Name Active I/O Note Pin Signal Active I/O ST1ON 23 ST1ON Stirrer 1 Scan ST1SCAN 24 ST1SCAN 5 Stirrer 1 Clear ST1CLS 25 ST1CLS 6 Stirrer 1 Busy 0 Sig ST1BUSY0 26 ST1BUSY0 7 Stirrer 1 Busy 1 Sig ST1BUSY1 27 ST1BUSY1 8 Stirrer 2 On ST2ON 28 ST2ON 9 Stirrer 2 Scan STS2CAN 29 STS2CAN 10 Stirrer 2 Clear ST2CLS 30 ST2CLS 11 Stirrer 2 Busy 0 Sig ST2BUSY0 31 ST2BUSY0 12 Stirrer 2 Busy 1 Sig ST2BUSY1 32 ST2BUSY1 13 NC 14 NC 15 +5V +5V IN VCC 16 +5V +5V IN VCC 33 +5V IN VCC Pin Signal Active I/O 1 IAAHOME H I Stirrer Driver CNN27>CNN3 (ISE Arm) Pin Signal 1 +5V 2 GND 3 Rotat Home Signal Name Active I/O Note ISE Arm IAAHOME H I 4 +5V 5 GND 6 U/D Home 7 +5V 8 GND 9 U/D Pitch 10 NC IAEHOME H I 2 IAEHOME H I IAEM1R H I 3 IAEM1R H I Pin Signal Active I/O CNN28>CNN3 (HCW Pump and IREF Cup Sensor) Pin Signal 1 +5V 2 GND 3 Home 4 +5V 5 GND 6 Pitch 7 +5V 8 GND 9 IREF 10 NC Signal Name Active I/O Note HCW Pump ISOHOME H I 28 ISOHOME H I ISOPITCH H I 29 ISOPITCH H I 73 ESMCCSSIREFCUP H I Pin Signal Active I/O IREF CUP Sensor ESMCCSSIREFCUP H I CNN29>CNN2 (Dust Cover Open/Close Sensor) Pin Signal 1 +5V 2 GND 3 Front Cover Sensr 4 NC 5 +5V 6 GND 7 Rear Cover Sensr 8 NC 9 NC 10 NC Signal Name Active I/O Note Dust Cover Open/Close Sensor FRONTCVROK H I 60 FRONTCVROK H I BACKCVROK H I 61 BACKCVROK H I Active I/O Pin Signal Active I/O 36 ISLUILED1(RCLED) H O CNN30>CNN2 (Reagent Supply Switches) Pin Signal 1 +5V 2 Rotat SW On Signal Name Note Reagent Supply SW ISLUILED1(RCLED) H O 3 +5V 4 Opn/Cls SW ON 5 GND 6 Rotat SW 7 GND 8 Opn/Cls SW 9 NC 10 NC ISLUISLED2(RCLED) H O 38 ISLUISLED1(RCLED) H O ISSW(RCSW) H I 91 ISSW(RCSW) H I ISLSW(RCSW) H I 92 ISLSW(RCSW) H I Pin Signal Active I/O CNN31>CNN2 and CNN4 (Automatic Reagent Loading) Pin Signal 1 +12V 2 12VGND 3 12V Snsr Si 4 4 +12V 5 12VGND 6 12V Snsr Si 7 +12V 8 12VGND 9 12V Snsr Si 2 10 +12V 11 12VGND 12 12V Snsr Si 1 13 +5V 14 GND 15 Sensor Si 27 16 +5V 17 GND 18 Sensor Si 26 19 +5V 20 GND 21 Sensor Si 25 22 +5V 23 GND 24 Sensor Si 24 25 +5V Signal Name Active I/O Note Cable Automatic Reagent Loading CNN4 RALOBJECT4 H I 32 RALOBJECT4 H I RALOBJECT3 H I 31 RALOBJECT3 H I RALOBJECT2 H I 30 RALOBJECT2 H I RALOBJECT1 H I 29 RALOBJECT1 H I SM13DLMT H I 28 SM13DLMT H I SM13PITCHOK H I 27 SM13PITCHOK H I SM13HOME H I 26 SM13HOME H I RALOPTION2 H I 25 RALOPTION2 H I 26 GND 27 Sensor Si 2 28 +5V 29 GND 30 Sensor Si 22 31 +5V 32 GND 33 Sensor Sig 21 34 +5V 35 GND 36 Sensor Sig 20 37 +5V 38 GND 39 Sensor Sig 19 40 +5V 41 GND 42 Sensor Sig 1 43 +5V 44 GND 45 Sensor Sig 17 46 +5V 47 GND 48 Sensor Sig 16 49 +5V 50 GND 51 Sensor Sig 15 52 +5V 53 GND 54 Sensor Sig 14 55 +5V 56 GND 57 Sensor Sig 13 58 NC 59 NC 60 NC 61 NC 62 NC 63 NC RALOPTION1 H I 24 RALOPTION1 H I SM12PITCHOK H I 23 SM12PITCHOK H I SM12HOME H I 22 SM12HOME H I RALGSNS3 H I 21 RALGSNS3 H I RALGSNS2 H I 20 RALGSNS2 H I RALGSNS1 H I 19 RALGSNS1 H I RALOPEN2 H I 18 RALOPEN2 H I RALOPEN1 H I 17 RALOPEN1 H I AC1UPHOME H I 33 AC1UPHOME H I AC1DOWNMOVE H I 34 AC1DOWNMOVE H I 11 RALACDOWNOK H I CNN2 RALACDOWNOK H I CNN4 64 +5V 65 GND 66 Sensor Sig 12 67 +5V 68 GND 69 Sensor Sig 11 70 +5V 71 GND 72 Sensor Sig 10 73 +5V 74 GND 75 Sensor Sig 9 76 +5V 77 GND 78 Sensor Sig 8 79 +5V 80 GND 81 Sensor Sig 7 82 +5V 83 GND 84 Sensor Sig 6 85 +5V 86 GND 87 Sensor Sig 5 88 +5V 89 GND 90 Sensor Sig 4 91 +5V 92 GND 93 Sensor Sig 3 94 +5V 95 GND 96 Sensor Sig 2 97 +5V 98 GND 99 Sensor Sig 1 SM4HOME H I 3 SM4HOME H I SM4PITCH H I 4 SM4PITCH H I SM4DLMT H I 5 SM4DLMT H I SM5HOME H I 6 SM5HOME H I SM5PITCH H I 7 SM5PITCH H I SM5DLMT H I 8 SM5DLMT H I SM6HOME H I 9 SM6HOME H I SM6PITCH H I 10 SM6PITCH H I SM6DLMT H I 11 SM6DLMT H I SM8HOME H I 14 SM8HOME H I SM8PITCH H I 15 SM8PITCH H I SM8DLMT H I 16 SM8DLMT H I 100 NC CNN32<CNN4 and CNN34 (AC/DC Driver Command) Pin Signal Signal Name Active I/O Note Cable Pin Signal Active I/O 1 Wash Elevator WS UP L O AC/DC Driver Command CNN34 1 WS UP L O 2 WS DOWN L O 2 WS DOWN L O 3 WS BK L O 3 WS BK L O STRZ UP L O 4 STRZ UP L O 5 STRZ DOWN L O 5 STRZ DOWN L O 6 STRZ BK L O 6 STRZ BK L O PWP UP L O 7 PWP UP L O 8 PWP DOWN L O 8 PWP DOWN L O 9 PWP BK L O 9 PWP BK L O AUXAC1UP L O 13 AUXAC1UP L O 11 AUXAC1DOWN L O 14 AUXAC1DOWN L O 12 AUXAC1BK L O 15 AUXAC1BK L O WCP ON L O 18 WCP ON L O WCP BK L O 19 WCP BK L O ICTP ON L O 20 ICTP ON L O ICTP BK L O 21 ICTP BK L O ICTR ON L O 22 ICTR ON L O ICTR BK L O 23 ICTR BK L O DTGP ON L O 16 DTGP ON L O DTGP BK L O 17 DTGP BK L O 4 7 10 Stirrer Arm Probe Wash Pump Reagent Supply Door Opn/Cls 13 NC 14 Cell Wash Pump 15 16 ISE Suction Pump 17 18 ISE Ref Pump 19 20 Detergent Pump 21 22 23 Cell Wash Vacuum Pump VP2 L O 26 VP2 L O 24 Cell Wash Dryer Pump VP1 L O 27 VP1 L O 25 Outer Wall Wash Pump WP MP L O 29 WP MP L O 26 Auto Power On MC1 L O 30 MC1 L O 28 Sampling Pump Valve ISPV L O 42 ISPV L O 29 R1 Pump Valve IR1PV L O 43 IR1PV L O 27 30 R2 Pump Valve IR2PV L O 44 IR2PV L O 31 ISE Suction Pinch Valve IAV L O 32 IAV L O 32 Detergent Valve 1 DTG1V L O 39 DTG1V L O 33 Detergent Valve 2 DTG2V L O 40 DTG2V L O 34 Cell Wash Valve CWV L O 38 CWV L O 35 S Probe Outer Wash Valve ESPV L O 33 ESPV L O 36 R Probe Outer Wash Valve ERPV L O 34 ERPV L O 37 STR Wash Valve SWV L O 35 SWV L O 38 DIW Supply Source Valve WBSV L O 36 WBSV L O 39 DIW Supply Pump PWV L O 37 PWV L O 40 ICT-Level Sensor IREF L O 49 IREF L O 41 Buzzer BUZZER L O 48 BUZZER L O 42 CRA CRA L O 50 CRA L O 43 DC Spare 1 AUXDC1 L O 67 AUXDC1 L O 44 DC Spare 2 AUXDC2 L O 68 AUXDC2 L O 45 NC 46 NC 47 AC Feedback Signal VP2 FB L I 62 VP2 FB L I 48 VP1 FB L I 50 VP1 FB L I 49 WPMP FB L I 64 WPMP FB L I 50 MC1 FB L I 65 MC1 FB L I ISPV FB L I 51 ISPV FB L I IR1PV FB L I 52 IR1PV FB L I 54 IR2PV FB L I 53 IR2PV FB L I 55 IAV FB L I 41 IAV FB L I 56 DTG1V FB L I 48 DTG1V FB L I 57 DTG2V FB L I 49 DTG2V FB L I 58 CWV FB L I 47 CWV FB L I 59 ESPV FB L I 42 ESPV FB L I 60 ERPV FB L I 43 ERPV FB L I 61 SWV FB L I 44 SWV FB L I 62 WBSV FB L I 45 WBSV FB L I 63 PWV FB L I 46 PWV FB L I 51 NC 52 DC Feedback Signal 53 CNN4 64 IREF FB L I 58 IREF FB L I 65 BUZZER FB L I 57 BUZZER FB L I 66 CRA FB L I 54 CRA FB L I 67 AUXDC1 FB L I 55 AUXDC1 FB L I 68 AUXDC2 FB L I 56 AUXDC2 FB L I I 60 FB GND 51 GND ACDC GND ACDC 52 GND ACDC +5V ACDC 55 +5V ACDC +5V ACDC 56 +5V ACDC 78 +5V ACDC 57 +5V ACDC 79 +5V ACDC 58 +5V ACDC 80 +5V ACDC 59 +5V ACDC 69 GND 70 NC 71 AC/DC DRIVER GROUND 72 73 NC 74 NC 75 NC 76 AC/DC DRIVER +5V 77 FB GND GND ACDC CNN34 I CNN33 < CNN35 and CNN36 (SMD Command) Pin Signal Signal Name Active I/O Note Cable Pin Signal Active I/O 1 Cuvette Rotate RCCW L O SMD Command CNN36 51 RCCW L O RCCCW L O 52 RCCCW L O RICCW L O 53 RICCW L O RICCCW L O 54 RICCCW L O ROCCW L O 55 ROCCW L O ROCCCW L O 56 ROCCCW L O SAZCW L O 63 SAZCW L O SAZCCW L O 64 SAZCCW L O SAHCW L O 61 SAHCW L O SAHCCW L O 62 SAHCCW L O R1ZINA L O 66 R1ZINA L O R1ZINB L O 67 R1ZINB L O 13 R1ZTDA L O 68 R1ZTDA L O 14 R1ZTDB L O 69 R1ZTDB L O 15 R1ZHOLD L O 70 R1ZHOLD L O R1HINA L O 61 R1HINA L O R1HINB L O 62 R1HINB L O 18 R1HTDA L O 63 R1HTDA L O 19 R1HTDB L O 64 R1HTDB L O 20 RR1HHOLD L O 65 RR1HHOLD L O R2ZINA L O 6 R2ZINA L O R2ZINB L O 7 R2ZINB L O 2 3 4 5 6 7 8 9 10 11 12 16 17 21 22 Reagent Inner Rotate Reagent Outer Rotate S Arm Up/Dwn S Arm Rotate R1 Arm Up/Dwn R1 Arm Rotate R2 Arm Up/Dwn CNN35 CNN36 23 R2ZTDA L O 8 R2ZTDA L O 24 R2ZTDB L O 9 R2ZTDB L O 25 R2ZHOLD L O 10 R2ZHOLD L O R2HINA L O 1 R2HINA L O R2HINB L O 2 R2HINB L O 28 R2HTDA L O 3 R2HTDA L O 29 R2HTDB L O 4 R2HTDB L O 30 R2HHOLD L O 5 R2HHOLD L O STR1HINA L O 11 STR1HINA L O STR1HINB L O 12 STR1HINB L O 33 STR1HTDA L O 13 STR1HTDA L O 34 STR1HTDB L O 14 STR1HTDB L O 35 STR1HHOLD L O 15 STR1HHOLD L O STR2HINA L O 16 STR2HINA L O STR2HINB L O 17 STR2HINB L O 38 STR2HTDA L O 18 STR2HTDA L O 39 STR2HTDB L O 19 STR2HTDB L O 40 STR2HHOLD L O 20 STR2HHOLD L O IAZINA L O 6 IAZINA L O IAZINB L O 7 IAZINB L O 43 IAZTDA L O 8 IAZTDA L O 44 IAZTDB L O 9 IAZTDB L O 45 IAZHOLD L O 10 IAZHOLD L O IAHINA L O 1 IAHINA L O IAHINB L O 2 IAHINB L O 48 IAHTDA L O 3 IAHTDA L O 49 IAHTDB L O 4 IAHTDB L O 50 IAHHOLD L O 5 IAHHOLD L O SPINA L O 31 SPINA L O SPINB L O 32 SPINB L O 53 SPTDA L O 33 SPTDA L O 54 SPTDB L O 34 SPTDB L O 55 SPHOLD L O 35 SPHOLD L O R1PINA L O 41 R1PINA L O 57 R1PINB L O 42 R1PINB L O 58 R1PTDA L O 43 R1PTDA L O 59 R1PTDB L O 44 R1PTDB L O 60 R1PHOLD L O 45 R1PHOLD L O R2PINA L O 51 R2PINA L O 62 R2PINB L O 52 R2PINB L O 63 R2PTDA L O 53 R2PTDA L O 64 R2PTDB L O 54 R2PTDB L O 65 R2PHOLD L O 55 R2PHOLD L O PPHCINA L O 36 PPHCINA L O 26 27 31 32 36 37 41 42 46 47 51 52 56 61 66 R2 Arm Rotate Stirrer 1 Rotate Stirrer 2 Rotate ISE Arm Up/Down ISE Arm Rotate Sampling Pump R1 Pump R2 Pump HCW Pump CNN35 CNN36 67 PPHCINB L O 37 PPHCINB L O 68 PPHCTDA L O 38 PPHCTDA L O 69 PPHCTDB L O 39 PPHCTDB L O 70 PPHCHOLD L O 40 PPHCHOLD L O AUTO1INA L O 36 AUTO1INA L O AUTO1INB L O 37 AUTO1INB L O 73 AUTO1TDA L O 38 AUTO1TDA L O 74 AUTO1TDB L O 39 AUTO1TDB L O 75 AUTO1HOLD L O 40 AUTO1HOLD L O AUTO2INA L O 46 AUTO2INA L O AUTO2INB L O 47 AUTO2INB L O 78 AUTO2TDA L O 48 AUTO2TDA L O 79 AUTO2TDB L O 49 AUTO2TDB L O 80 AUTO2HOLD L O 50 AUTO2HOLD L O AUTO3INA L O 56 AUTO3INA L O AUTO3INB L O 57 AUTO3INB L O 83 AUTO3TDA L O 58 AUTO3TDA L O 84 AUTO3TDB L O 59 AUTO3TDB L O 85 AUTO3HOLD L O 60 AUTO3HOLD L O AUXINA L O 46 AUXINA L O 87 AUXINB L O 47 AUXINB L O 88 AUXTDA L O 48 AUXTDA L O 89 AUXTDB L O 49 AUXTDB L O 90 AUXHOLD L O 50 AUXHOLD L O 71 GND SMD 72 GND SMD 71 GND SMD 72 GND SMD +5V SMD 79 +5V SMD 98 +5V SMD 80 +5V SMD 99 +5V SMD 79 +5V SMD 100 +5V SMD 80 +5V SMD Active I/O 71 72 76 77 81 82 86 91 Automatic Reagent Loading 1 Automatic Reagent Loading 2 Automatic Reagent Loading 3 Spare SMD GND GND SMD 93 GND SMD 94 GND SMD NC 96 NC 97 SMD +5V CNN36 GND SMD 92 95 CNN35 CNN35 CNN36 CNN35 CNN36 CNN38>CNN37 (Flow Sensors Signal) Pin Signal 1 +5V 2 GND 3 Spl Line Flow Signal 4 +5V 5 GND Signal Name Active I/O Note Pin Signal Flow Sensor Signal SPFS I SPFS I 6 Rgt Line Flow Signal 7 +5V 8 GND 9 Str Line Flow Signal 10 NC RPFS I RPFS I STFS I STFS I ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. ICT Pre Amp Board The ICT Pre-Amp Board gain is factory set and does not require adjustment. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. IREF Level Detect Board Correct LED 1 and LED 2 Status Before the ICT assembly drives down and aspirates ICT reference solution from the cup, LED 1 Read Status and LED 2 ICT Reference Cup Full turn ON. LED 2 turns on only when the c4000 module CPU is reading the status of the level sense electrodes in the ICT Reference Cup. During this brief status check (when LED 2 is on) observe LED 1. If the cup is full during the status check, LED 1 will be on at the same time as LED 2. If the cup is empty, LED 1 will be off when LED 2 is on. 1. Just before the ICT assembly moves down to aspirate, both LED2 and LED1 turn RED momentarily (cup is full), then turn off. 2. After the ICT assembly aspirates and moves up, LED2 turns RED momentarily and then turns off. LED1 never turns on because the IREF Cup is empty. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. LLS Board LLS Block Diagram TLLS PWB Test Point Observation Point LLS Detection Voltage Shortage Sample Voltage TP:A5 >0.50v – – >0.50v >0.30v – – >0.40v >0.30v – – >0.40v TP:B5 TP:T5 Sample Pipettor Liquid Level Sense Note Stop signal for sample Confirmation for sample short Stop signal for reagent 1 Confirmation for reagent 1 short Stop signal for reagent 2 Confirmation for reagent 2 short Reagent Pipettor Level Sense ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Thermal Bath Controller (TBC) Board The TBC Board maintains water bath temperature at 37+ 0.3C, keeping the reaction liquid temperature within +0.1C target value. Ambient air temperature sensors connect to this board. The board provides 24 and 100VAC to control heater, pump, fan, and motor valve. The pump fan and motor valve are controlled by a photocoupler. The heater is controlled by the CPU using the thyristor (solid-state semi conductor acting as a switch to turn the heater On/Off). Thermal Bath Controller Board (side view) Thermal Bath Controller Board (connectors) Thermal Bath Control New Digital TBC LEDs, Switches and Test pins Switches No. Name SW1 Function MODE / PARAMETER (4bits DIP-switch) MODE (Temperature adjustment) Bit1 Bit2 Mode OFF OFF Disable: (Default) OFF ON Disable ON OFF PC mode (For the future) It can be adjusted by PC ON ON Manual mode: It can be adjusted by SW3 and SW4. When Bit1 and Bit2 are turned "ON", temperature can be adjusted by SW3 and SW4. And the adjusted temperature is stored on this board. When the temperature adjustment was completed, Bit1 and Bit2 must be turned "OFF". Temperature can't be adjusted by SW3 and SW4. Temperature is controlled to become the adjusted temperature. PARAMETER (For the future.) Bit3 Bit4 Parameter OFF OFF Type A (Default) OFF ON Type B ON OFF Type C ON ON Type D SW3 UP (Push Button switch) Temperature adjustment up button: Approximately 0.02°C/Step SW4 Down (Push Button switch) Temperature adjustment down button: Approximately 0.02°C/Step LEDs No. Name Color Function LED1 HEATER RED ON: The heater is ON. LED2 CPU CMD Red ON: The heater is ON signal is received. LED3 UP RED ON: SW3 (UP) is pushed. LED4 DOWN RED ON: SW4 (DOWN) is pushed. LED5 POWER RED ON: +5V is supplied. Test Pins No. Test Point TP1 +5V TP2 GND TP3 Thermistor signal TP4 Signal to detect zero cross point of AC 100V for phase control ARCHITECT c4000 Service and Support Manual (Version 204733-113) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Power Supply Links Main Power Supply Transformer ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Main Power Supply Outside voltage is supplied to an isolation transformer via circuit protector and line filter. The transformer converts input voltages to 24VAC and 100VAC. Primary Input The c4000 uses external AC voltage from 200VAC to 240VAC (± 10%). Five corresponding selector plugs; 200, 208, 220, 230, and 240 are available on the transformer. The plug closest to external voltage should be selected. Output The power supply generates continuous 100VAC to individual power supply. The individual power supply generates: • +3.3, +5 and ±12VDC • +12VDC separately for magnetic values • +24VDC for the SMD • +5, ±12 and ±36VDC for the RSH • ±15VDC for analog (feedback) circuitry • +11.45VDC lamp power supply The powers supply is equipped with no fuse breakers (NFB), an emergency stop switch, and power indicator LED board for monitoring active status of various DC power supplies. Relays control AC system output and low-voltage supply. Power-INDI Board The power indicator board monitors the output state of the DC System Power Supply. Each voltage is monitored and the LED illuminates when respective voltage is present. The state of the LED can be monitored from the window of the Power Supply. Main Power Supply (front view) Connector Function 1 Fan 2 Circuit Protector 3 Voltage Monitor LEDs 4 Analyzer Switch 5 RSH Switch 6 External Power 7 Main Power Switch 8 Individual Power Supply Adjustment Constant Voltage Power Supply Specifications PS Output Usage Voltage (V) Current (A) 100VAC 6A AC Motors, Pumps, Heaters 24VAC 1A Stirrers Drive, Temperature Control Circuit 1 +3.3VDC 8A Card Cage (motherboard internal power supply) 1 +5VDC 15A Card Cage (control logic), CNBD, TBC 1 +/-12VDC 2A Card Cage (motherboard internal power supply) 2 +/-15VDC 1A Analog Power Supply (STRD. Card Cage) 3 +12VDC 8A Magnetic Valve Drive, CPUM, +12V, CNBD 4 +11.6VDC 2A Lamp, CPUM 5 +12VDC 27A UC, Thermostat, Rgt Sply Cntr, Fans 6 +24VDC 18A Stepping Motor Drive (+24V2, +24V1), CPUM 7 +36VDC 18A J1, J2 (RSH Drive Power Supply) 8 +5VDC 10A J5, J6 (RSH CPU Control) 8 +12VDC 2A J3, J4 RSH Communications 8 -12VDC 1A J3, J4 RSH Communications Power Supply Voltage Status LEDs LED Voltage Assembly Voltage Monitored Source 1 +24V Stepping Motor Drive (+24V2, +24V1), CPUM PS6 2 +36V J1, J2 (RSH Drive Power Supply) PS7 3 +12V UC, Thermostat, Rgt Sply Cntr, Fans PS5 4 +12V Card Cage (motherboard internal power supply) PS1 5 -12V Card Cage (motherboard internal power supply) PS1 6 +11.5 Lamp, CPUM PS4 7 +12V Magnetic Valve Drive, CPUM, +12V, CNBD PS3 8 +15V Analog Power Supply (STRD. Card Cage) PS2 9 -15V Analog Power Supply (STRD. Card Cage) PS2 10 +5V Card Cage (control logic), CNBD, TBC PS1 11 +5V J5, J6 (RSH CPU Control) PS8 12 +12V Card Cage (motherboard internal power supply) PS1 13 12V 14 +3.3V Power Supply Variable Resistors Circuit Protection PS1 Card Cage (motherboard internal power supply) PS1 Breaker Rating (A) Control AC2 3A 24VAC line for the TBC and STR-DRIVER AC3 6A 100VAC line for the AC/DC DRIVER AC4 6A 100VAC line for the TBC AC5 6A 100VAC line for the Cooling Fan CP1 8A +24VDC SMD Line 1 CP2 8A +24VDC SMD Line 2 CP3 8A +24VDC SMD Line 3 CP4 8A +24VDC SMD Line 4 CP5 8A +24VDC SMD Line 5 CP6 8A +24VDC SMD Line 6 CP7 4A +24VDC SMD Line 7 Main Power Supply (rear connector plate) Power Supply Connector Pin Outs Power Supply Connections Connector Function J3 RSH +12VDC Output CPUM Output for monitoring DC Voltage AUTO Switched 100VAC input for lower voltage power supply c+RSH c4000 Standalone RSH Connector i+RSH c4000 Integration RSH Connector J4 RSH +/-12VDC Output R/F Front Power Switch Thermostat +12VDC Output ON/OFF Spare 100VAC Output RSH Input Side Connector for the RSH Power Supply J1 RSH +36VDC Output CNBD +5VDC, +12VDC for the CNN Board TBC 24VAC and +5VDC for Water Bath Control MC IN Input Side for the Magnet Contact MC OUT Output Side for the Magnet Contact J2 RSH +36VDC Output UC +12VDC for Unit Cooling AC100V AC100V Output STRD 24VAC and +/-15VDC Output for the STR-Driver 2 +12V +12VDC for the AC/DC Driver +24V1 SMD +24VDC Output (1) AC 100VAC Output for the AC/DC Driver, TBC and Cooling Fan J5 RSH +5VDC Output CPU +5VDC, +/-15VDC, +3.3VDC, +/-12VDC Output for the card cage chassis +24V2 SMD +24VDC Output (2) Transformer 100VAC, 24VAC input for the secondary side of the transformer J6 RSH +5VDC Output LAMP +11.5VDC output for the Lamp Lamp 1 +11.5V 2 GND CPU 1 +5V 2 GND 3 +15V 4 +5V 5 GND 6 AGND 7 +5V 8 GND 9 GND 10 +3.3V 11 GND 12 NC 13 +12V 14 GND 15 -12V +24 V2 1 +24V 2 +24V 3 +24V 4 GND 5 GND 6 GND 7 +24V 8 +24V 9 +24V 10 GND 11 GND 12 GND 13 +24V 14 GND 15 NC +24 V2 1 +24V 2 +24V 3 +24V 4 GND 5 GND 6 GND 7 +24V 8 +24V 9 +24V 10 GND 11 GND 12 GND UC 1 +12V 2 +12V 3 +12V 4 GND 5 GND 6 GND 7 +12V 8 GND 9 NC J6 1 +5V 2 +5V 3 +5V 4 +5V 5 GND 6 GND 7 GND 8 GND 9 NC J1 1 +36V 2 +36V 3 +36V 4 GND 5 GND 6 GND J2 1 +36V 2 +36V 3 +36V 4 GND 5 GND 6 GND CPUM 1 +24V 2 +12V 3 +11.5V 4 24GND 5 12GND 6 11.5GND +12V 1 +12V 2 GND 3 NC 4 NC 5 NC J5 1 +5V 2 +5V 3 GND 4 GND CNBD 1 +5V 2 5GND 3 +12V 4 12GND J3 1 +12V 2 -12V 3 GND 4 GND J4 1 +12V 2 +12V 3 -12V 4 -12V 5 GND 6 GND Transformer 1 AC100V MC1 T3 2 AC100V MC1 T3 3 AC100V MC1 T1 4 AC100V MC1 T4 5 AC100V MC1 T3 6 AC100V MC1 T2 7 AC24V MC1 T5 8 NC 9 AC24V MC1 T6 MC IN 1 AC24V MC IN 2 AC100V MC IN 3 AC100V RSH MC IN 4 AC24V MC IN 5 AC100V MC IN 6 AC100V RSH MC IN MC OUT 1 AC24V MC OUT 2 AC100V MC OUT 3 AC100V RSH MC OUT 4 AC24V MC OUT 5 AC100V MC OUT 6 AC100V RSH MC OUT AUTO 1 MC1 A1 2 MC1 A2 3 NC R/F 1 BK A1 2 NC 3 TB3 10B Thermostat 1 TB1 7B 2 PS5 L AC100V 1 TB1 7B (AC100V) 2 TB1 5B (AC100V) TBC 1 PS1 +5V15A 2 NC 3 TB2 10A (AC24V) 4 PS1 +5V GND 5 NC 6 TB2 9A (AC24V) STRD 1 TB2 10A (AC24V) 2 TB2 9A (AC24V) 3 PS2 +15V 1A 4 PS2 +/-15V GND 5 PS2 –15V1A AC 1 AC100V AC/DC Driver 2 AC100V AC/DC Driver 3 AC100V TBC 4 AC100V TBC 5 AC100V FANS 6 AC100V FANS 7 NC 8 NC 9 NC ON/OFF 1 RSH Power Switch IN (AC100V) 2 RSH Power Switch IN (AC100V) 3 NC RSH 1 TB2 4B (AC100V) 2 TB2 2B (AC100V) c-iRSH 1 TB1 2B (AC100V) 2 TB1 1B (AC100V) i-iRSH 1 TB1 4B (AC100V) 2 TB1 3B (AC100V) Power Supply Schematics Schematic Diagram Link Power Supply Diagram 7A_9004.pdf Page 1 Power Supply Diagram 7A_9004.pdf Page 2 ARCHITECT c4000 Service and Support Manual (Version 204733-107) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Transformer • Input - 200, 208, 220, 230, and 240VAC (± 10%) at 50Hz/60Hz ± 170Hz • Output - 24VAC and 100VAC (± 10%) at 50Hz/60Hz ± 1%Hz Magnetic Contactor (Relay) The contactor is internal to the power supply. MC1 • Input - Receives 24VAC and 100VAC input from the transformer • Output - Provides 24VAC for TBC and STRDRV - Provides 100VAC to 3 terminal blocks for distribution to power supplies PS1, PS2, PS3, PS4, PS5, PS6, PS7, PS8 Power Transformer Cable Pinout Pin Voltage Current Variance Output Wire Gauge 1-6 100VAC 20A ±3% 16AWG, 600V 7, 9 24VAC 1A ±3% 18AWG, 600V Circuit Protector The 24VAC output line is protected by a 4 amp circuit protector (CPI) on front of the transformer. The 24VAC and 100VAC output lines are protected by thermal fuses. ARCHITECT c4000 Service and Support Manual (Version 204733-107) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. C1.01 Optics Assembly Version - 204738-101_4873_1 List/Part Numbers List/Part Number Description 7-205181-01 OPTICS ASSY 7-205181-02 OPTICS ASSY Inspect tools for damage, ensure calibration is not expired and replace if necessary. C1.01 Optics Assembly Time Required 35 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Caution Hot Surface Removal Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the right rear cover. Remove the rear right deck cover [K]. Remove the Lamp housing cover [G]. 2. 3. 4. Remove Source Lamp 1. 2. Remove the thumbscrew [1] securing the source lamp holder and remove the lamp holder. Lay the holder out of the way. Note There is no need to disconnect lamp from terminal block. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Optics Assembly 1. 2. 3. 4. Follow the cables from the Fan and Pre-Amp board and disconnect them (both, the shutter cable SHUT-SHAT and fan PHOTO-FAN). Remove the Pre-Amp cover [2] by loosening the left screw and removing the right screw with the ground cable [1]. Disconnect the OPT cable from the Pre-Amp board. While supporting the optics assembly from the bottom, remove the four (4) screws [3] securing the optics and remove it from the instrument. Replacement Action Steps Install Optics Assembly 1. 2. 3. 4. 5. 6. Reference While supporting the optics assembly from the bottom, secure it to the processing module frame using four (4) screws [3]. Connect the pre-amp cable [4]. Install the Pre-Amp cover and ground wires, route the cables between the gasket material (of the cover) and the housing. Insert the ground cable and secure the cover with the screws. Note Ensure the cover slides into the guides in the bottom of the optics housing. The fit between the guides in the optic housing and the pre-amp cable cover is tight. Connect the shutter cable (SHUT to SHAT). Connect the fan (PHOTO-FAN) cable. Install Source Lamp 1. Install source lamp holder in optics assembly and tighten the screw. Prepare for Operation 1. 2. 3. 4. Install the lamp housing cover [G]. Install the rear right deck cover [K]. Install right rear cover. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. 5. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 1001 - Optics Trigger Sensor Check 2 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C1.02 Optics Shutter Version - 204738-101_4874_2 List/Part Numbers List/Part Number Description 7-204228-01 Optics Shutter 7-204228-02 Optics Shutter, new style 7-35005695-01 Shutter, Universal Inspect tools for damage, ensure calibration is not expired and replace if necessary. C1.02 Optics Shutter Time Required 50 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Caution Hot Surface Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover. Remove the rear right deck cover [K]. Remove the Lamp housing cover [G]. 2. 3. 4. Remove Source Lamp 1. 2. Remove the thumbscrew [1] securing the source lamp holder and remove the lamp holder. Lay the holder out of the way. Note There is no need to disconnect lamp from terminal block. Remove Optics Assembly 1. 2. 3. 4. Follow the cables from the Fan and Pre-Amp board and disconnect them (both, the shutter cable SHUT-SHAT and fan PHOTO-FAN). Remove the Pre-Amp cover [2] by loosening the left screw and removing the right screw with the ground cable [1]. Disconnect the OPT cable from the Pre-Amp board. While supporting the optics assembly from the bottom, remove the four (4) screws securing the optics and remove it from the instrument [3]. Remove Optics Housing Cover 1. Remove the five (5) screws securing the optics cover to the optics. Caution Do not try and clean any components inside of the optics assembly. Damage can occur by touching the lens, the grating, or the diode array. Caution Do not try to adjust or unscrew any screw with red screw lock on them. Remove Cable Clamp 1. Unscrew the screw securing the cable clamp for the shutter cable [1]. Pull the shutter cable up through slot behind the preamp board. Cut the wire tie on the cable clamp and save the cable clamp and screw. 2. 3. Remove Shutter Remove Universal Shutter From Mount Plate 1. Remove the two (2) screws [1] securing the shutter mounting bracket [2] and remove the shutter and bracket from the optics assembly. These instructions are for removing the new 735005695-01 Universal Shutter. If you are not removing the Universal Shutter, proceed to action step: Remove Shutter From Shutter Mount. 1. 2. 3. Remove the (2) two screws [1] securing the shutter to the shutter mounting plate. Remove shutter [2] from the mounting plate [3]. Proceed to action step: Install Universal Shutter On Mount Plate. Remove Shutter From Shutter Mount 1. 2. Remove the three (3) screws [1] securing the shutter to the shutter mount. Remove shutter from the shutter mount. [2]. Note the position of the shutter while removing. Replacement Action Steps Install Universal Shutter On Mount Plate These instructions are for the new 7-35005695-01 Universal Shutter. If you are not installing the Universal Shutter, proceed to action step: Install Shutter On Mount. 1. Place the new universal shutter [1] on the new mount plate [2] and secure with the (2) two screws [3]. Note The Universal Shutter Kit contains the mount plates needed to install the shutter on the c4000, c16000 and c8000 (two different versions) Optics Assemblies. The unused mount plates are discarded. Reference Install Universal Shutter On Mounting Post 1. 2. Place the shutter, with shutter mount, on the mounting post and secure with the two (2) screws [1]. Proceed to action step: Install Cable Clamp. Note Apply pressure to the shutter mount in the direction of the lamp housing while tightening the two (2) screws. Install Shutter On Shutter Mount 1. Place the shutter on the shutter mount and secure with three (3) screws. Note The shutter will only fit on the shutter mount in one position. Install Shutter 1. Place the shutter, with shutter mount, on the mounting post and secure with two (2) screws. Note Install Cable Clamp 1. 2. 3. Apply pressure to the shutter mount in the direction of the lamp housing while tightening the two (2) screws. Secure the shutter cable to the cable clamp with a wire tie. Secure the cable clamp [1] to the optics assembly with one screw. Route the shutter cable behind the lens mount and then through the slot behind the preamp board. Install Optics Housing Cover 1. Install the optics cover to the optics housing and secure with five (5) screws. Install Optics Assembly 1. While supporting the optics assembly from the bottom, secure it to the processing module frame using four (4) screws. Connect the pre-amp cable [4]. Install the Pre-Amp cover and ground wires, route the cables between the gasket material (of the cover) and the housing. Insert the ground cable and secure the cover with the screws. Note Ensure the cover slides into the guides in the bottom of the optics housing. The fit between the guides in the optic housing and the preamp cable cover is tight. Connect the shutter cable (SHUT to SHAT). Connect the fan (PHOTO-FAN) cable. 2. 3. 4. 5. 6. Install Source Lamp 1. Install source lamp holder in optics assembly and tighten the screw. Prepare for Operation 1. 2. 3. 4. Install the lamp housing cover [G]. Install the rear right deck cover [K]. Install right rear cover. Power ON the processing module. Note The system control center power For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 5. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 1001 - Optics Trigger Sensor Check 2 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C1.03 Fan, Optics Assembly Version - 204738-101_4875_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. C1.03 Fan, Optics Assembly Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Caution Hot Surface Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the right rear cover. Remove the rear right deck cover [K]. Remove the Lamp housing cover [G]. 2. 3. 4. Remove Source Lamp 1. 2. Remove the thumbscrew [1] securing the source lamp holder and remove the lamp holder. Lay the holder out of the way. Note There is no need to disconnect lamp from terminal block. Remove Optics Assembly 1. 2. 3. 4. Follow the cables from the Fan and Pre-Amp board and disconnect them (both, the shutter cable SHUT-SHAT and fan PHOTO-FAN). Remove the Pre-Amp cover [2] by loosening the left screw and removing the right screw with the ground cable [1]. Disconnect the OPT cable from the Pre-Amp board. While supporting the optics assembly from the bottom, remove the four (4) screws [3] securing the optics and remove it from the instrument. Remove Optics Fan Note 1. 2. 3. Fan airflow direction is OUT of the housing. Remove the four (4) screws securing the preamp cable cover mount. [1] Remove the four (4) screws [2] securing the fan to the optics assembly. Note The screw hardware is not captured and there is a flat washer and a split washer on each screw. Remove the Optics Fan. Replacement Action Steps Install Optics Fan Note 1. 2. Install Optics Assembly Reference 1. 2. 3. 4. Verify fan airflow direction is OUT of the housing. Install the fan to the optics assembly using four (4) screws [2], each with a flat washer, and a split washer. Note Air flow direction is out of the housing. Install the pre-amp cable cover mount. While supporting the optics assembly from the bottom, secure it to the processing module frame using four (4) screws. Connect the pre-amp cable [4]. Install the Pre-Amp cover and ground wires, route the cables between the gasket material (of the cover) and the housing. Insert the ground cable and secure the cover with the screws. Note Ensure the cover slides into the guides in the bottom of the optics housing. The fit between the guides in the optic 5. 6. housing and the pre-amp cable cover is tight. Connect the shutter cable (SHUT to SHAT). Connect the fan (PHOTO-FAN) cable. Install Source Lamp 1. Install source lamp holder in optics assembly and tighten the screw. Prepare for Operation 1. 2. 3. 4. Install the lamp housing cover [G]. Install the rear right deck cover [K]. Install right rear cover. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. 5. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. ARCHITECT c4000 Service and Support Manual (Version 204733-107) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Verification Procedures Order VP Description VP Detail / Note 1 G14 - Fan Verify that the fan is rotating and that the air flows in the correct direction. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. null Expected Results C1.05 Filter, Heat Absorbing Version - 204738-101_4876_1 List/Part Numbers List/Part Number Description 7-202484-01 Optics Heat Filter Inspect tools for damage, ensure calibration is not expired and replace if necessary. C1.05 Filter, Heat Absorbing Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Caution Hot Surface Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Note The heat absorbing filter should never be removed with the processing module powered on. Remove the lamp housing cover [G]. 2. Remove Duct 1. Remove Heat Absorbing Filter Removal and replacement steps are left to the discretion of the Abbott trained repair person. Note 1. 2. 3. Do not touch the heat absorbing filter; oil from your hands can damage the filter. Pull the heat filter mount plate [1] out of the optics assembly. Loosen the screws securing each filter holder and move the holder to the sides [2]. Remove the heat filter [3] from the mount. Replacement Action Steps Install Heat Absorbing Filter Note 1. 2. 3. Prepare for Operation Reference 1. 2. 3. Do not touch the heat absorbing filter, as the oil from your hands can damage the filter. Install the heat filter [3] on the mount. [1] Place back the filter holder [2] and secure them with their screws. Install the filter mount plate [1] in the optics assembly with the heat filter holders toward the optics housing. Install the Lamp housing cover. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C2.01 Motor Assy, Reaction Carousel Version - 204738-101_4877_1 List/Part Numbers List/Part Number Description 7-205184-01 MOTOR ASSY., REACTION CRSL. Inspect tools for damage, ensure calibration is not expired and replace if necessary. C2.01 Motor Assy, Reaction Carousel Time Required 35 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the cuvette segments from the reaction carousel and store in a safe place. Remove the lamp housing cover [G]. Remove center deck cover-center left cover [D]. Remove Rear deck-Rear right cover [K]. Remove the left rear cover. 2. 3. 4. 5. 6. Remove Reaction Carousel Motor Assembly 1. 2. 3. 4. Remove the two (2) screws securing the carousel gear cover and move the cover to the side. [1] Disconnect the RD-M and RD-S plugs from the connector plate. Disconnect RD Pitch from the motor. Remove the cable ties. Remove Reaction Carousel Motor Assembly (continued) 1. While holding motor (so it does not fall), remove the three (3) screws securing the motor assembly. [2] Replacement Action Steps Install Reaction Carousel Motor Assembly 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. 2. Reference Install the motor assembly to upper frame securing the three (3) screws. Install the carousel gear cover. Connect RD Pitch to the motor. Connect the RD-M and RD-S plug to the connector plate. Install the lamp housing cover [G]. Install center deck cover-center left cover [D]. Install Rear deck-Rear right cover [K]. Power ON the processing module. Note To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Replace the left rear cover. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 1001 - Optics Trigger Sensor Check 3 1008 - Optics Total Test 4 1120 - Sample Pipettor Calibration 5 1121 - R1 Pipettor Calibration 6 1122 - R2 Pipettor Calibration 2 1009 - Beam Height Alignment 9 P-618 ICT Alignment 7 P-620 Mixer Alignment 8 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C2.02 Roller Assy, Cuvette Align Version - 204738-101_4878_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. C2.02 Roller Assy, Cuvette Align Time Required 25 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the lamp housing cover [G]. 2. Remove Roller Assembly 1. 2. Mark the position of the roller assembly in relationship to the upper processing module frame. Note This will ensure proper alignment of the roller in relationship to the gear when reinstalled. Remove the two (2) screws securing the assembly to the frame and remove from the processing module. [1] Replacement Action Steps Install Roller Assembly 1. 2. 3. Install the assembly on the frame using the marks made during removal. Secure the assembly to the frame using two (2) screws. Manually rotate the reaction carousel main gear Reference 4. Prepare for Operation 1. 2. 3. 4. and make sure the wheel turns constantly. Adjust the roller assembly until the wheel rotates smoothly. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform 1009 Beam Height Alignment. Install the lamp housing cover [G]. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 1120 - Sample Pipettor Calibration 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C2.03 Wheel, Common, Roller Assy Version - 204738-101_4879_1 List/Part Numbers List/Part Number Description 7-205311-01 WHEEL, COMMON, ROLLER ASSY Inspect tools for damage, ensure calibration is not expired and replace if necessary. C2.03 Wheel, Common, Roller Assy Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the lamp housing cover [G]. 2. Remove Roller Assembly 1. 2. Remove Roller 1. 2. Mark the position of the roller assembly in relationship to the upper processing module frame. Note This ensures proper alignment of the roller in relationship to the gear when reinstalled. Remove the two (2) screws securing the assembly to the frame and remove from the processing module. [1] Remove the one screw securing the roller to the post. Disassemble the roller assembly into the spacer, wheel, bearing, and shaft. Replacement Action Steps Install Roller 1. 2. Install Roller Assembly 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. Reference Assemble the roller. The spacer is closest to the post, followed by the wheel, bearing and shaft. Install the one screw to attach the assembled roller to the post. Position the assembly on the frame using the marks made during removal. Note If installing the assembly closest to the motor gear, ensure the assembly is seated properly on the guide pin. Install the assembly to the frame using two (2) screws. Manually rotate the reaction carousel main gear and make sure the wheel turns constantly. Adjust the roller assembly until the wheel rotates smoothly. Tighten the two (2) screws. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Install the lamp housing cover [G]. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 3 1121 - R1 Pipettor Calibration 4 1122 - R2 Pipettor Calibration 1 1009 - Beam Height Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C2.04 Guide Assembly Version - 204738-101_4880_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. C2.04 Guide Assembly Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the lamp housing cover [G]. 2. Remove Guide Assembly 1. 2. Mark the position of the guide assembly in relationship to the upper processing module frame. Note This will ensure proper alignment of the wheel in relationship to the gear when reinstalled. Remove the two (2) screws securing the assembly to the frame and remove from the processing module. [1] Replacement Action Steps Install Guide Assembly 1. 2. Position the assembly on the frame using the marks made during removal. Install the assembly to the frame using two (2) screws. Reference 3. 4. Prepare for Operation 1. 2. 3. Manually rotate the reaction carousel main gear and make sure the wheel turns constantly. Adjust the guide assembly until the wheel rotates smoothly. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Install the lamp housing cover [G]. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 3 1121 - R1 Pipettor Calibration 4 1122 - R2 Pipettor Calibration 1 1009 - Beam Height Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C2.05 Wheel with Bearing, Common, Guide Assy Version - 204738-101_4881_1 List/Part Numbers List/Part Number Description 7-202490-01 Wheel & bearing, com guide Inspect tools for damage, ensure calibration is not expired and replace if necessary. C2.05 Wheel with Bearing, Common, Guide Assy Time Required 20 min Tools/Materials Standard Tool Kit (Used on the Guide Assembly and Spring Tensioned Guide Assembly.) WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the lamp housing cover [G]. 2. Remove Guide Assembly 1. 2. Remove Wheel with Bearing 1. 2. Mark the position of the guide assembly in relationship to the upper processing module frame. Note This will ensure proper alignment of the wheel in relationship to the gear when reinstalled. Remove the two (2) screws securing the assembly to the frame and remove from the processing module. [1] Remove one screw securing the wheel with bearing from the post. On the spring tensioned guide assembly there is a nut securing the screw in place. Remove the nut. Replacement Action Steps Install Wheel Bearing 1. 2. Install Guide Assembly 1. 2. 3. 4. Prepare for Operation 1. 2. 3. Reference Install the wheel with bearing to the post using one screw. Install the one screw with nut, if this is the spring tensioned guide assembly. Position the assembly on the frame using the marks made during removal. Install the assembly to the frame using two (2) screws. Manually rotate the reaction carousel main gear and make sure the wheel turns constantly. Adjust the guide assembly until the wheel rotates smoothly. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Install the lamp housing cover [G]. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 3 1121 - R1 Pipettor Calibration 4 1122 - R2 Pipettor Calibration 1 1009 - Beam Height Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. C2.06 Motor Coupling, Reaction Carousel Version - 204738-101_5039_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. C2.06 Motor Coupling, Reaction Carousel Time Required 50 min Tools/Materials Standard Tool Kit While performing this procedure, if you are exposed to reagents, calibrators or controls, the following statements may apply: WARNING Potential Biohazard WARNING Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Remove Reaction Carousel Motor Assembly. (C2.01 Motor Assy, Reaction Carousel) Remove Motor Coupling 1. Loosen the two (2) upper set screws [1] securing the coupling to the drive gear shaft. Remove the four (4) screws securing the motor to the motor frame. Loosen the lower two (2) set screws [2] securing the coupling to the motor shaft and remove coupling. 2. 3. Replacement Action Steps Install Motor Coupling 1. 2. 3. 4. 5. 6. Install coupling onto the gear drive shaft. Align set screws with the flat sides of the shaft. Tighten the set screws to the gear drive shaft. Install the motor and insert shaft into the coupling. Secure the motor to the motor frame using the four (4) screws. Align the set screws on the Reference coupling with the flat sides of the motor shaft and tighten set screws. Install Reaction Carousel Motor Assembly 1. Install Reaction Carousel Motor Assembly. (C2.01 Motor Assy, Reaction Carousel) Verification Procedures Order VP Description 1 1001 - Optics Trigger Sensor Check 3 1008 - Optics Total Test 4 1120 - Sample Pipettor Calibration 5 1121 - R1 Pipettor Calibration 6 1122 - R2 Pipettor Calibration VP Detail / Note Expected Results 2 1009 - Beam Height Alignment 9 P-618 ICT Alignment 7 P-620 Mixer Alignment 8 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-109) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. D1.01 Reagent Supply Center Version - 204738-101_4882_3 List/Part Numbers List/Part Number Description 7-205901-01 REAGENT CAROUSEL ASSEMBLY, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.01 Reagent Supply Center Time Required 90 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Caution Lifting Hazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. Preparation Open the reagent supply center cover. Remove the reagent trays; store in an appropriate place. Power OFF the c4000 processing module (use the main circuit breaker). Remove the top covers. Remove the right rear cover. Remove the left side covers. Remove the reagent probes. Remove the following deck covers: • Front left cover [A] • Front center cover [B] • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] Remove the reagent carousel motor cover [1]. 1. 2. 3. 4. 5. 6. 7. 8. 9. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation (continued) 1. 2. 3. Preparation (continued) 1. 2. 3. 4. Remove the green ground cable [2]. Disconnect [3]: • CVR-M • CVR-OP • CVR-CL Remove the three (3) screws securing the reagent carousel cover and remove the cover. Loosen the two (2) screws [4] on the reflective plate assembly to the RSH rail and remove plate. Remove the two (2) screws from the black left front panel to the top left side panel. Note Do not remove the black panel. Remove the top screw and loosen the bottom screw securing the CRA cable bracket to the top left side panel and slide off the bracket from the panel. Remove the four (4) Allen head screws that secure the top left side panel and remove the panel. Remove Reagent Supply Center 1. 2. 3. Remove the screw that secures the cover sensor and bracket assembly. Disconnect the following cables: a. At the front [1]: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear [2]: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Remove assembly from the module. Remove Reagent Supply Center (continued) 1. 2. 3. 4. Disconnect the drain tubing [1] from the LCW drain manifold. Clamp tubing end to prevent leaking from the reagent carousel. Remove the four (4) screws securing the reagent supply center to the mounting locations. Lift the reagent supply center from the processing module. Note Do not pinch cables or bend brackets when lifting the carousel. Note You may need assistance lifting the carousel. Replacement Action Steps Install Reagent Supply Center 1. 2. 3. 4. 5. Place the reagent supply center into the processing module. Align the attachment holes on the carousel with the alignment pin [2] and screw hole [1] on the processing module. Note You may need assistance lifting the reagent supply center. Tighten the four (4) screws. Connect the drain tubing into the LCW drain manifold. Connect the following cables: a. At the front: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear: Reference 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. Install 1. Reagent Supply Center (continued) 2. 3. Prepare for 1. Operation 2. - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Install the cover sensor and bracket assembly and secure with the screw. Install the top left side panel and secure with four (4) screws. Attach the black left front panel to the top left side panel with two (2) screws. Install the reflective plate and secure with two (2) screws. Slide the CRA cable bracket onto the two (2) screws on the top left side panel and tighten the screws. Install the reagent carousel cover using three (3) screws. Connect: • CVR-M • CVR-OP • CVR-CL Connect the green ground cable to the frame. Install the reagent carousel motor cover. Install the left side covers. Install the right rear cover. Install the Top Deck covers: • Front left cover [A] • Front center cover [B] • Center left cover [D] • Rear center left cover [I] • Rear left cover [H] Install the Reagent Probes. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Install the reagent cartridges in the reagent supply center. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 3525 - Temperature Status 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration 4 G30 - Reagent Bar Code Label Scan N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results Perform a SCAN F4 under the REAGENT Icon. Verify all reagent bar code labels are read. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT�ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. D1.02 TED, Reagent Supply Center Version - 204738-101_4883_3 List/Part Numbers List/Part Number Description 7-205216-01 TED, RGT. CRSL. Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.02 TED, Reagent Supply Center Time Required 30 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Caution Lifting Hazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. Preparation Open the reagent supply center cover. Remove the reagent trays; store in an appropriate place. Power OFF the c4000 processing module (use the main circuit breaker). Remove the top covers. Remove the right rear cover. Remove the left side covers. Remove the reagent probes. Remove the following deck covers: • Front left cover [A] • Front center cover [B] • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] Remove the reagent carousel motor cover [1]. 1. 2. 3. 4. 5. 6. 7. 8. 9. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation (continued) 1. 2. 3. Preparation (continued) 1. 2. 3. 4. Remove the green ground cable [2]. Disconnect [3]: • CVR-M • CVR-OP • CVR-CL Remove the three (3) screws securing the reagent carousel cover, and remove the cover. Loosen the two (2) screws securing the reflective plate assembly [4] to the RSH rail and remove plate. Remove the two (2) screws securing the black left front panel to the top left side panel. Note Do not remove the black panel. Remove the top screw and loosen the bottom screw securing the CRA cable bracket to the top left side panel and slide off the bracket from the panel. Remove the four (4) Allen head screws that secure the top left side panel and remove the panel. Remove Reagent Supply Center 1. 2. 3. Remove the screw that secures the cover sensor and bracket assembly. Disconnect the following cables: a. At the front [1]: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear [2]: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Remove assembly from the module. Remove Reagent Supply Center (continued) 1. 2. 3. 4. Remove TED 1. 2. Disconnect the drain tubing [1] from the LCW drain manifold. Clamp tubing end to prevent leaking from the reagent carousel. Remove the four (4) screws securing the reagent supply center to the mounting locations. Lift the reagent supply center from the processing module. Note Do not pinch cables or bend brackets when lifting the carousel. Note You may need assistance lifting the carousel. Remove all segment bases to allow access to the reagent carousel. Note When removing the bases for the exterior ring, leave the barcode reader gap [1] in place for reference during reassembly. Locate the external ring carousel motor and remove the two (2) screws [2] holding the bearing across from the carousel motor, lift it and remove the external dented carousel. Remove TED (continued) 1. 2. Remove the four (4) screws [4] from the corresponding TED. Turn the reagent supply center upside down to access the TED fan assembly and covers. Note If needed, remove the TED fan assembly by removing the two (2) screws [5] in the perimeter of the reagent supply center and sliding the TED fan assembly. Remove TED (continued) 1. 2. 3. Loosen the two (2) screws [6] close to the center of the Reagent Supply center. Remove the two (2) screws [7] on the side of the TED cover. Disconnect the corresponding UC Cable and remove the TED. Replacement Action Steps Install TED 1. 2. 3. 4. 5. 6. 7. 8. Clean the base [1] where the TED sits. Note Make sure the receptacle is clean to apply heat sink compound. Place the TED on the base. Place the cover and secure with the two (2) screws. Connect the UC cable. Note If removed, replace the TED fan assembly, and secure with the two (2) screws. Turn the reagent supply center upside down. Secure the TED with the screws; alternate tightening for even contact. Place the external ring in position. Make sure it seats properly on all bearings Place the bearing across from the carousel motor on the guide pin [2], tighten the screws. Make sure the dented carousel ring moves evenly. Reference Install TED 1. (continued) 2. Install Reagent Supply Center 1. 2. 3. 4. 5. Assemble the internal reagent segment holder. Note The internal reagent segment holder has a pin guide [3] for correct position. Replace all reagent segment bases for the external carousel. Seat properly and in appropriate order. Place the reagent supply center in the processing module. Align the attachment holes on the carousel with the alignment pin [2] and screw hole [1] on the processing module. Note You may need assistance lifting the reagent supply center. Secure with four (4) screws. Connect the drain tubing into the LCW drain manifold. Connect the following cables: a. At the front: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear: - ROCM - RICM - THERMO - RCS 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Install the cover sensor and bracket assembly and secure with the screw. Install the top left side panel and secure with four (4) screws. Attach the black left front panel to the top left side panel with two (2) screws. Slide the CRA cable bracket onto the two (2) screws on the top left side panel and tighten the screws. Install the reagent supply center cover using three (3) screws. Connect: • CVR-M • CVR-OP • CVR-CL Connect the green ground cable to the frame. Install the reagent carousel cover using three (3) screws. Install the left side cover. Install the right rear cover. Install the Top Deck covers: • Front Deck: Front left cover [A] • Front Deck: Front center cover [B] • Center Deck: Center left cover [D] • Rear Deck: Rear center left cover [I] • Rear Deck: Rear left cover [H] Install 1. Reagent 2. Supply Center (continued) Install the Reagent Probes. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Prepare for 1. Operation 2. Install the reagent cartridges in the reagent supply center. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 3525 - Temperature Status 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration 4 G30 - Reagent Bar Code Label Scan N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Perform a SCAN F4 under the REAGENT Icon. Verify all reagent bar code labels are read. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT�ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. D1.03 2D Reagent Barcode Reader Version - 204738-101_4884_3 List/Part Numbers List/Part Number Description 7-206380-01 BARCODE READER, REAGENT, C4000 7-206380-02 BARCODE READER, REAGENT, C4000 Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.03 2D Reagent Barcode Reader Time Required 25 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. Preparation Power OFF the c4000 processing module (use the main circuit breaker). Remove the front left cover [A]. 1. 2. Remove Barcode Reader Assembly 1. 2. Disconnect the bottle sensor and barcode reader cables [1]. Unscrew the two (2) screws [2] securing the barcode reader and bottle sensor bracket assembly and remove assembly. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Barcode Reader Assembly (continued) 1. Remove the two (2) screws [3] from the barcode reader frame. Unplug the barcode reader cable from the barcode reader. 2. Replacement Action Steps Install Barcode 1. 2. Install the barcode reader to the frame with two (2) screws. Plug the barcode reader cable in the barcode reader. Reference Reader 3. Assembly 4. 5. Install the barcode reader and bottle sensor bracket assembly using two (2) screws. Connect the bottle sensor and barcode reader cables. Install the front left cover. Prepare 1. for Operation 2. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Select BarCode Readers category from the Diagnostics screen under the System icon. 3. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 2 G30 - Reagent Bar Code Label Scan Perform a SCAN F4 under the REAGENT Icon. Verify all reagent bar code labels are read. 1 P-606 Reagent Supply Center Bar Code Reader Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT�ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. D1.04 Inner Carousel Motor, Reagent Supply Center Version - 204738-101_4885_3 List/Part Numbers List/Part Number Description 7-205185-02 MOTOR CRSL., INNER Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.04 Inner Carousel Motor, Reagent Supply Center Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Caution Lifting Hazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. Preparation Open the reagent supply center cover. Remove the reagent trays; store in an appropriate place. Power OFF the c4000 processing module (use the main circuit breaker). Remove the top covers. Remove the right rear cover. Remove the left side covers. Remove the reagent probes. Remove the following deck covers: • Front left cover [A] • Front center cover [B] • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] Remove the reagent carousel motor cover [1]. 1. 2. 3. 4. 5. 6. 7. 8. 9. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation (continued) 1. 2. 3. Preparation (continued) 1. 2. 3. 4. Remove the green ground cable [2]. Disconnect [3]: • CVR-M • CVR-OP • CVR-CL Remove the three (3) screws securing the reagent carousel cover, and remove the cover. Loosen the two (2) screws [4] on the reflective plate assembly to the RSH rail and remove plate. Remove the two (2) screws from the black left front panel to the top left side panel. Note Do not remove the black panel. Remove the top screw and loosen the bottom screw securing the CRA cable bracket to the top left side panel and slide off the bracket from the panel. Remove the four (4) Allen head screws that secure the top left side panel and remove the panel. Remove Reagent Supply Center 1. 2. 3. Remove the screw that secures the cover sensor and bracket assembly. Disconnect the following cables: a. At the front [1]: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear [2]: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Remove assembly from the module. Remove Reagent Supply Center (continued) 1. 2. 3. 4. Remove Motor 1. 2. 3. 4. 5. Disconnect the drain tubing [1] from the LCW drain manifold. Clamp tubing end to prevent leaking from the reagent carousel. Remove the four (4) screws securing the reagent supply center to the mounting locations. Lift the reagent supply center from the processing module. Note Do not pinch cables or bend brackets when lifting the carousel. Note You may need assistance lifting the carousel. Loosen the top two (2) Allen screws [1] on the motor shaft to the coupler. Remove the four (4) screws used to mount the motor to the bracket. Lift the motor ensuring it disengages from the coupler. Follow the motor cable and disconnect it from the bracket. Release cable clamps. Replacement Action Steps Install Motor 1. 2. 3. 4. 5. Install Reagent Supply Center 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. Install 1. Reagent 2. Supply 3. Center (continued) 4. 5. Install the motor onto the bracket, guiding the motor shaft into the coupler. Note The flat section of the shaft faces the Allen screws. Secure the motor to the bracket using screws. Secure the coupler to the shaft. Route the cable connector to the bracket and insert the cable into the frame. Place the cable into cable clamps and close the clamps. Place the reagent supply center in the processing module. Align the attachment holes on the carousel with the alignment pin [2] and screw hole [1] on the processing module. Note You may need assistance lifting the reagent supply center. Secure with the four (4) screws. Connect the drain tubing into the LCW drain manifold. Connect the following cables: a. At the front: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Install the cover sensor and bracket assembly and secure with the screw. Install the top left side panel and secure with four (4) screws. Attach the black left front panel to the top left side panel with two (2) screws. Slide the CRA cable bracket onto the two (2) screws on the top left side panel and tighten the screws. Install the reagent supply center cover using three (3) screws. Connect: • CVR-M • CVR-OP • CVR-CL Connect the green ground cable to the frame. Install the reagent motor cover using one (1) screw. Install the left side cover. Install the right rear cover. Install the Top Deck covers: • Front Deck: Front left cover (A) • Front Deck: Front center cover (B) • Center Deck: Center left cover (D) • Rear Deck: Rear center left cover (I) • Rear Deck: Rear left cover (H) Install the Reagent Probes. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to Reference ensure proper initialization. Prepare for 1. Operation 2. Install the reagent cartridges in the reagent supply center. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 3525 - Temperature Status 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration 4 G30 - Reagent Bar Code Label Scan N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results Perform a SCAN F4 under the REAGENT Icon. Verify all reagent bar code labels are read. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT�ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. D1.05 Outer Carousel Motor, Reagent Supply Center Version - 204738-101_4886_3 List/Part Numbers List/Part Number Description 7-205186-01 MOTOR CRSL., OUTER 7-205186-02 MOTOR CRSL., OUTER Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.05 Outer Carousel Motor, Reagent Supply Center Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Caution Lifting Hazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. Preparation 1. 2. 3. The processing module must be in Stopped or Offline status. Open the reagent supply center cover. Remove the reagent trays; store in an appropriate place. Power OFF the c4000 processing module (use the main circuit breaker). 4. Remove the top covers. 5. Remove the two (2) captive screws securing the right rear cover, and remove the cover. 6. Remove the left side covers. 7. Remove the reagent probes. 8. Remove the following deck covers: • Front left cover [A] • Front center cover [B] • Center left cover [D] • Rear center left cover [I] • Rear left cover [H] 9. Remove the reagent carousel motor cover [1]. 10. Remove the green ground cable [2]. 11. Disconnect [3]: Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. • CVR-M • CVR-OP • CVR-CL 12. Remove the three (3) screws securing the reagent carousel cover and remove the cover. Preparation (continued) 1. 2. 3. 4. Loosen the two (2) screws [4] on the reflective plate assembly to the RSH rail and remove plate. Remove the two (2) screws from the black left front panel to the top left side panel. Note Do not remove the black panel. Remove the top screw and loosen the bottom screw securing the CRA cable bracket to the top left side panel and slide off the bracket from the panel. Remove the four (4) Allen head screws that secure the top left side panel and remove the panel. Remove Reagent Supply Center 1. 2. 3. Remove the screw that secures the cover sensor and bracket assembly. Disconnect the following cables: a. At the front [1]: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear [2]: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Remove assembly from the module. Remove Reagent Supply Center (continued) 1. 2. 3. 4. Remove Motor 1. 2. 3. 4. 5. Disconnect the drain tubing [1] from the LCW drain manifold. Clamp tubing end to prevent leaking from the reagent carousel. Remove the four (4) screws securing the reagent supply center to the mounting locations. Lift the reagent supply center from the processing module. Note Do not pinch cables or bend brackets when lifting the carousel. Note You may need assistance lifting the carousel. Loosen the top two (2) Allen screws [1] on the motor shaft to the coupler. Remove the four (4) screws used to mount the motor to the bracket. Lift the motor ensuring it disengages from the coupler. Follow the motor cable and disconnect it from the bracket. Release cable clamps. Replacement Action Steps Install Motor 1. 2. 3. 4. 5. Install Reagent Supply Center 1. 2. 3. 4. 5. 6. 7. 8. 9. Install 1. Reagent Supply 2. Center (continued) 3. 4. 5. 6. 7. 8. 9. Prepare for 1. Operation 2. Reference Install the motor onto the bracket, guiding the motor shaft into the coupler. Note The flat section of the shaft faces the Allen screws. Secure the motor to the bracket using screws. Secure the coupler to the shaft. Route the cable connector to the bracket and insert the cable into the frame. Place the cable into cable clamps and close the clamps. Place the reagent supply center in the processing module. Align the attachment holes on the carousel with the alignment pin [2] and screw hole [1] on the processing module. Note You may need assistance lifting the reagent supply center. Secure with four (4) screws. Connect the drain tubing into the LCW drain manifold. Connect the following cables: a. At the front: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Install the cover sensor and bracket assembly and secure with the screw. Install the top left side panel and secure with four (4) screws. Attach the black left front panel to the top left side panel with two (2) screws. Slide the CRA cable bracket onto the two (2) screws on the top left side panel and tighten the screws. Install the reagent supply center cover using three (3) screws. Connect: • CVR-M • CVR-OP • CVR-CL Connect the green ground cable to the frame. Install the reagent motor cover using one (1) screw. Install the left side cover. Install the right rear cover. Install the Top Deck covers: • Front Deck: Front left cover [A] • Front Deck: Front center cover [B] • Center Deck: Center left cover [D] • Rear Deck: Rear center left cover [I] • Rear Deck: Rear left cover [H] Install the Reagent Probes. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module. Install the reagent cartridges in the reagent supply center. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. < Verification Procedures Order VP Description 1 3525 - Temperature Status 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration 4 G30 - Reagent Bar Code Label Scan N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results Perform a SCAN F4 under the REAGENT Icon. Verify all reagent bar code labels are read. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT�ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. D1.06 Bottom Fan, Reagent Supply Center Version - 204738-101_4887_4 List/Part Numbers List/Part Number Description 7-205187-01 FAN, COOLING, RGT., EXTERNAL Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.06 Bottom Fan, Reagent Supply Center Time Required 90 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Caution Lifting Hazard Caution Chemical Hazard Removal(continued) Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. Preparation Open the reagent supply center cover. Remove the reagent trays; store in an appropriate place. Power OFF the c4000 processing module (use the main circuit breaker). Remove the top covers. Remove the right rear cover. Remove the left side covers. Remove the reagent probes. Remove the following deck covers: • Front left cover [A] • Front center cover [B] • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] Remove the reagent carousel motor cover [1]. 1. 2. 3. 4. 5. 6. 7. 8. 9. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation (continued) 1. 2. 3. Preparation (continued) 1. 2. 3. 4. Remove the green ground cable [2]. Disconnect [3]: • CVR-M • CVR-OP • CVR-CL Remove the three (3) screws securing the reagent carousel cover and remove the cover. Loosen the two (2) screws [4] on the reflective plate assembly to the RSH rail and remove plate. Remove the two (2) screws from the black left front panel to the top left side panel. Note Do not remove the black panel. Remove the top screw and loosen the bottom screw securing the CRA cable bracket to the top left side panel and slide off the bracket from the panel. Remove the four (4) Allen head screws that secure the top left side panel and remove the panel. Remove Reagent Supply Center 1. 2. 3. Remove the screw that secures the cover sensor and bracket assembly. Disconnect the following cables: a. At the front [1]: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear [2]: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Remove assembly from the module. Remove Reagent Supply Center (continued) 1. 2. 3. 4. Remove Bottom Fan, Reagent Supply Center Disconnect the drain tubing [1] from the LCW drain manifold. Clamp tubing end to prevent leaking from the reagent carousel. Remove the four (4) screws securing the reagent supply center to the mounting locations. Lift the reagent supply center from the processing module. Note Do not pinch cables or bend brackets when lifting the carousel. Note You may need assistance lifting the carousel. Note 1. 2. 3. Fan airflow direction is OUT at the back of the module. Place the reagent supply center upside-down on a table. Remove the two (2) screws [1] from the fan on the reagent supply center. Release the cable from the clamps and bracket [2] and remove the fan. Replacement Action Install Bottom Fan, Reagent Supply Center Install Reagent Supply Center Steps Reference Note Verify fan airflow direction is OUT at the back of the module. 1. 2. Install the fan and secure it with the screws. Insert the cable in the clamp and the bracket. 1. Place the reagent supply center in the processing module. Align the attachment holes on the carousel with the alignment pin [2] and screw hole [1] on the processing module. Note You may need assistance lifting the reagent supply center. Secure with four (4) screws. Connect the drain tubing into the LCW drain manifold. Connect the following cables: a. At the front: - BWF - Bar Code Reader (RBCR) - Tube Sensor (BOT SNSR) b. At the rear: - ROCM - RICM - THERMO - RCS - Fan Cable - UC1 - UC2 - UC3 - UC4 - UC FAN cables. c. At the inner side between the wash cups: - CVR Sensor cable Install the cover sensor and bracket assembly and secure with the screw. Install the top left side panel and secure with four (4) screws. Attach the black left front panel to the top left side panel with two (2) screws. Slide the CRA cable bracket onto the two (2) screws on the top left side panel and tighten the screws. Install the reagent supply center cover using three (3) screws. Connect: • CVR-M • CVR-OP • CVR-CL Connect the green ground cable to the frame. Install the reagent motor cover using one (1) screw. Install the left side cover. Install the right rear cover. Install the Top Deck covers: • Front Deck: Front left cover [A] • Front Deck: Front center cover [B] • Center Deck: Center left cover [D] • Rear Deck: Rear center left cover [I] • Rear Deck: Rear left cover [H] 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. Install 1. Reagent 2. Supply Center (continued) Install the Reagent Probes. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Prepare for 1. Operation Install the reagent cartridges in the reagent supply center. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating 2. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. instructions, Subsection: System startup, pause, and shutdown. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT�ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Verification Procedures Order VP Description 1 3525 - Temperature Status 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration 4 G30 - Reagent Bar Code Label Scan N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. null VP Detail / Note Perform a SCAN F4 under the REAGENT Icon. Verify all reagent bar code labels are read. Expected Results D1.07 Internal Fan, Air Cir, Reagent Supply Center Version - 204738-101_4888_4 List/Part Numbers List/Part Number Description 7-205188-01 FAN, COOLING, RGT., INTERNAL Inspect tools for damage, ensure calibration is not expired and replace if necessary. D1.07 Internal Fan, Air Cir, Reagent Supply Center Time Required 60 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Probe Stick Hazard WARNING Chemical Hazard Removal Action Steps Prerequisite 1. Preparation 1. 2. The processing module must be in Stopped or Offline status. Open the reagent supply center cover. Remove the reagent trays, store in an appropriate place. 3. Power OFF the c4000 processing module (use the main circuit breaker). 4. Remove the Reagent Probes. 5. Remove the right rear cover 6. Remove the following: • Y-Shaped reagent supply center maintenance cover. • Front left cover [A] • Rear left cover [H] 7. Remove the screw securing reagent cover motor cover. 8. Remove the green ground cable [2]. 9. Disconnect [3]: • CVR-M • CVR-OP • CVR-CL 10. Remove the three (3) screws securing the reagent carousel cover, and remove the cover. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation (continued) 1. Remove Internal Fan Remove all segment bases to allow access to the reagent carousel. Note When removing the bases for the exterior ring, leave the barcode reader gap [1] in place for reference during reassembly. Note 1. 2. 3. 4. Fan airflow direction is UP into the reagent supply center. Disconnect FAN cable and open the cable clamp securing the cable. Open the cable clamps inside the reagent supply center used to secure the fan cable. Remove the screws used to secure the fan to the reagent supply center base and the screws holding the cable. Pull cable with connector through the insulation and remove the fan. Replacement Action Steps Install Internal Fan Note 1. 2. 3. 4. Install Reagent Supply Center Reference 1. Verify fan airflow direction is UP into the reagent supply center. Secure the fan with the screws. Route the cable through the cable clamps and secure it. Insert the connector through the insulation. Connect the FAN cable to the bracket. Assemble the internal reagent segment holder. Align with pin guide [3]. 2. Replace all reagent segment bases for the external carousel. Seat properly and in appropriate order. 3. Install the reagent supply center cover using three (3) screws. 4. Connect: • CVR-M • CVR-OP • CVR-CL 5. Connect the green ground cable to the frame. 6. Install the reagent motor cover using one (1) screw. 7. Install the right rear cover. 8. Install the Top Deck covers: • Front Deck: Front left cover (A) • Rear Deck: Rear left cover (H) • Y-Shaped reagent supply center maintenance cover. 9. Install the Reagent Probes. 10. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Prepare 1. for Operation 2. Install the reagent cartridges in the reagent supply center. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT�ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Verification Procedures Order VP Description 1 3525 - Temperature Status 2 1121 - R1 Pipettor Calibration 3 1122 - R2 Pipettor Calibration 4 G30 - Reagent Bar Code Label Scan VP Detail / Note Perform a SCAN F4 under the REAGENT Expected Results Icon. Verify all reagent bar code labels are read. N/A null G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. E1.01 PROBE Version - 96728-111_392_1 List/Part Numbers List/Part Number Description 08C94-42 Probe Inspect tools for damage, ensure calibration is not expired and replace if necessary. E1.01 PROBE Time Required Tools/Materials 20 min. #2 Phillips screw driver (optional) WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Prerequisite 1. The processing module must be in Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Warming, or Ready status. Preparation 1. Lift the appropriate processing center cover(s) to access the probe(s). 2. Locate the probe to be replaced on the appropriate pipettor. Note Refer to the processing center map for pipettor locations (R1, R2, S, or ST). 3. Verify the probe is over the wash station. For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. To move the reagent probe over the wash station, perform a Startup on the appropriate module, and go to Remove Probe. To move the sample probe over the wash station, go to Step4. 4. Select the Pipettors category from the diagnostic screen under the system icon. 5. Initiate diagnostic procedure 1160 Pipettor Move to move the sample probe over the wash station. Remove Probe 1. Place an absorbent towel under the probe tubing connection. 2. Loosen the metal fitting on the probe and remove the probe tubing from the probe. [1] 3. Carefully loosen the screw (captive) on top of the probe clamp until resistance is felt. [2] Note A #2 Phillips screwdriver can be used to loosen the screw. Note Loosening the captive screw beyond the point of resistance can cause the clip to fall apart. 4. Push up on the blue button located under the boom arm. [3] 5. Swing the probe free of the clip, lift the probe and remove it from the boom arm. [4] Replacement Action Install Probe Steps 1. Slide the probe into the boom arm. [1] 2. Swing the probe to the probe clip and snap it in place under the clip. [2] 3. Tighten the screw (captive) on top of the probe clamp. [3] Note A #2 Phillips screwdriver may be used to tighten the screw. Do not overtighten as stripping may result. 4. Connect the probe tubing to the Reference probe and finger tighten the probe fitting. [4] 5. To return the sample pipettor to the home position, complete diagnostic procedure 1160 Pipettor Move. Prepare for Operation For further information on performing maintenance procedures, refer to the ARCHITECT System Operations Manual, Section 9: Service and 1. Select As Needed category from the Maintenance, Subsection: As-needed maintenance procedures. maintenance screen under the system icon. 2. Perform maintenance procedure 2130 Flush Fluids, to remove any air that may be present. 3. Visually check for leaks while performing the flush. Note If leaks are seen, repeat the installation procedure. 4. Remove the absorbent towel. 5. Close the processing center cover(s). Verification Procedures Order VP Description 1 1111 - Sample Pipettor Calibration 4 1117 - STAT Pipettor Calibration 2 1112 - R1 Pipettor Calibration 3 1113 - R2 Pipettor Calibration N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E2.01 Wash Cup, Sample Version - 204738-101_4889_1 List/Part Numbers List/Part Number Description 7-93131-02 Wash Cup, Sample 7-93131-03 Wash Cup, Sample (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. E2.01 Wash Cup, Sample Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Probe Stick Hazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. 2. Remove the sample pipettor head cover. Loosen the screw for the probe ground wire, and disconnect the tubing from the probe. Remove the screw securing the sample probe while removing the probe. Remove the front center cover [B]. 3. 4. Remove Sample Wash Cup 1. 2. 3. Disconnect the tubing from the sides of the sample wash cup. [1] Remove the two (2) screws securing the sample wash cup to the mounting bracket. [2] Lift up the wash cup and remove the drain tubing from the bottom of the cup. Secure drain tubing to keep from dropping inside of instrument. [3] Replacement Action Steps Install Sample Wash 1. Cup 2. 3. 4. 5. 6. Connect the drain tubing to the bottom of the sample wash cup. Install the wash cup to the mounting bracket using the two (2) screws. Connect the two incoming tubes to the sides of the sample wash cup. Install the front center cover. Install the sample probe on the sample pipettor and secure using a screw. Connect the tubing and secure the ground Reference 7. Prepare for Operation 1. wire. Install the sample pipettor head cover. Perform P-608 Sample Wash Cup Position Alignment. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from Stopped to System Operations Manual, Section 5: Operating instructions, Ready, perform the Startup procedure. Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 3 1120 - Sample Pipettor Calibration 2 1175 - Wash Cup Test 1 P-608 Sample Wash Cup Position Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.02 Filter Housing Version - 204738-101_4893_1 List/Part Numbers List/Part Number Description 7-202499-01 Filter housing Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.02 Filter Housing Time Required Not Assessed Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Preparation 1. 1. 2. 3. Remove Filter Housing 1. 2. 3. Reference Drain the water bath For further information, refer to the ARCHITECT System Operations (System/Diagnostics/Modules/1101 Robotics Manual, Section 1: Use or function, Subsection: System status. Test Tool - Option 11). The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the rear left cover. Unscrew the threaded cap [3] at the bottom of the filter unit and drain any remaining water. Unscrew the filter housing [1] counterclockwise (CCW) from the lid. [2] If the filter needs to be replaced, discard the filter. Replacement Action Steps Install Filter Housing 1. 2. Screw the threaded cap [3] into the base of the filter housing. Install the filter support plate [4] on the nipple at the bottom of the filter housing. Reference Install Filter Housing (continued) 1. Install the filter onto the disk and screw the filter housing [1] clockwise (CW) onto the cap [2], making sure the hollow center of the filter aligns with the fingers inside the cap. [5] Remove Air from Filter Housing 1. Power ON the instrument and wait for the water to fill the filter housing. 2. 3. 4. Prepare for Operation 1. 5. Check for air bubbles at the tubing attached For information on the startup procedure, refer to the ARCHITECT to the degasser manifold and the tubing going System Operations Manual, Section 5: Operating instructions, to the probe wash pumps. Subsection: System startup, pause, and shutdown. Check for air bubbles in the tubing on both side of the 12 VDC CKD (DI Water Cuvette Wash) Valve. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. If air bubbles are observed, select the As Needed category from the Maintenance screen under the System icon. Perform 2132 Flush Water Lines. 1. Install the left rear cover. 2. 3. 4. Install Rear Covers Loosen the top right bolt (inlet side) on the cap to allow air to escape as the filter housing is filled with water. Make sure you remove any remaining air through the bolt. Note There should not be any bubbles in the filter. Tighten the bolt. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 G63 - Tubing / Fitting / Filter Verify tubing / fitting / filter is securely installed and does not obstruct instrument components. Ensure no leaks and/or bubbles are in the tubing / fitting / filter. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E2.03 Wash Cup, R1 & R2 Reagent Probes Version - 204738-101_4891_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E2.03 Wash Cup, R1 & R2 Reagent Probes Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Probe Stick Hazard Caution Chemical Hazard Removal Action Steps Prerequisite Note 1. Access Wash Cup 1. 2. 3. 4. Remove Reagent Wash Cup Reference 1. 2. 3. This procedure for Reagent Wash Cup R1A and R1B. The processing module must be in Stopped or Offline status. Remove the reagent pipettor head cover. Remove the screw for the probe ground wire and disconnect the tubing from the probe. Remove the screw securing the reagent probe while removing the probe. Remove the center deck-center deck cover. [D] Remove the screws securing the reagent wash cup to the upper frame. [1] Disconnect the tubing from the side of the reagent wash cup. [2] Lift up the wash cup and remove the drain tubing from the bottom of cup. Make sure to secure it and do not lose it. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Replacement Action Steps Install Reagent Wash Cup 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. Reference Insert the drain tubing into the wash cup. Connect the incoming water tubing to the side of the wash cup. Install the wash cup to the upper frame using the screws. Install the reagent probe on the reagent pipettor and secure using a screw. Connect tubing and secure ground wire. Perform P-610 Reagent Wash Cup (R1 and R2) Alignment accordingly. Install the center deck cover [D]. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 3 1121 - R1 Pipettor Calibration 4 1122 - R2 Pipettor Calibration 2 1175 - Wash Cup Test 1 P-610 Reagent Wash Cup (R1 and R2) Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.01 Pump, Water Bath Fill Version - 204738-101_4892_1 List/Part Numbers List/Part Number Description 7-202498-01 Pump, Water Bath Fill 7-207600-01 Pump, c4 Bath Fill With Cable Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.01 Pump, Water Bath Fill Time Required Not Assessed Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Preparation 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. 3. 4. Remove Water Bath Fill Pump 1. 2. 3. 4. 5. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Set the processing module must in Stopped or Offline status. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover. Remove the screw securing the tubing clamp to the frame. Remove the plastic peristaltic pump head by rotating the housing counter clockwise (CCW) one sixteenth (1/16) turn and then pulling outward on the housing to remove it from the pump. Remove the four screws securing the pump housing panel. Disconnect the WBSV connector. Disconnect the amp fuse connector. Remove Water Bath Fill Pump (continued) 1. From below the housing frame, squeeze both tabs on the pump to release the pump from the frame and remove the pump. [1] Replacement Action Steps Install Water Bath Fill Pump 1. 2. 3. 4. 5. 6. 7. 8. Prepare for Operation 1. Reference Snap in the pump. Insert the tabs on the housing into the corresponding notches on the pump face and rotate the housing clockwise until the pump housing does not rotate anymore. Connect the amp fuse connector. Connect the WBSV connector. Secure the pump housing panel to the frame using one screw. Install the plastic peristaltic pump head. Attach the tubing clamp to the frame using the one screw. Attach the rear left cover. Power ON the processing module. Note The system control center power must be on prior to For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 2. 3. 4. 5. turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2134 Change Water Bath. Check for leaks. Verification Procedures Order VP Description VP Detail / Note 1 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.02 Filter Housing Version - 204738-101_4893_1 List/Part Numbers List/Part Number Description 7-202499-01 Filter housing Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.02 Filter Housing Time Required Not Assessed Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Preparation 1. 1. 2. 3. Remove Filter Housing 1. 2. 3. Reference Drain the water bath For further information, refer to the ARCHITECT System Operations (System/Diagnostics/Modules/1101 Robotics Manual, Section 1: Use or function, Subsection: System status. Test Tool - Option 11). The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the rear left cover. Unscrew the threaded cap [3] at the bottom of the filter unit and drain any remaining water. Unscrew the filter housing [1] counterclockwise (CCW) from the lid. [2] If the filter needs to be replaced, discard the filter. Replacement Action Steps Install Filter Housing 1. 2. Screw the threaded cap [3] into the base of the filter housing. Install the filter support plate [4] on the nipple at the bottom of the filter housing. Reference Install Filter Housing (continued) 1. Install the filter onto the disk and screw the filter housing [1] clockwise (CW) onto the cap [2], making sure the hollow center of the filter aligns with the fingers inside the cap. [5] Remove Air from Filter Housing 1. Power ON the instrument and wait for the water to fill the filter housing. 2. 3. 4. Prepare for Operation 1. 5. Check for air bubbles at the tubing attached For information on the startup procedure, refer to the ARCHITECT to the degasser manifold and the tubing going System Operations Manual, Section 5: Operating instructions, to the probe wash pumps. Subsection: System startup, pause, and shutdown. Check for air bubbles in the tubing on both side of the 12 VDC CKD (DI Water Cuvette Wash) Valve. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. If air bubbles are observed, select the As Needed category from the Maintenance screen under the System icon. Perform 2132 Flush Water Lines. 1. Install the left rear cover. 2. 3. 4. Install Rear Covers Loosen the top right bolt (inlet side) on the cap to allow air to escape as the filter housing is filled with water. Make sure you remove any remaining air through the bolt. Note There should not be any bubbles in the filter. Tighten the bolt. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 G63 - Tubing / Fitting / Filter Verify tubing / fitting / filter is securely installed and does not obstruct instrument components. Ensure no leaks and/or bubbles are in the tubing / fitting / filter. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.03 Filter, Water Bath Version - 204738-101_4894_2 List/Part Numbers List/Part Number Description 7-202500-01 Filter, water bath 7-206374-01 BUFFER FILTER SUPPORT DISK Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.03 Filter, Water Bath Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required Not Assessed Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Chemical Hazard WARNING Splash/Spray Hazard Removal Action Steps Reference Preparation 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. 3. 4. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 10). The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the rear left cover. VIDEO Note Remove Filter Housing and Filter Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove 1. Filter Housing and 2. Filter 3. 4. Clamp the tubing on both ports of the filter unit. Unscrew the threaded cap at the bottom of the filter unit and drain any remaining water. Unscrew the filter housing [1] counterclockwise (CCW) from the lid. [2] If the filter needs to be replaced, discard the filter. Replacement Action Steps VIDEO Install Filter Housing and Filter Note Reference Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Filter Housing and Filter Note To help minimize bubbles, presoak the replacement filter—prior to installing—in a plastic bag of distilled water > 1 hour prior to filter installation. 1. 2. 3. Screw the threaded cap [3] into the base of the filter housing. Install the filter support plate [4] on the fitting at the bottom of the filter housing. Install the filter onto the disk and screw the filter housing [1] clockwise (CW) onto the cap [2], making sure the hollow center of the filter aligns with the fingers inside the cap. [5] Note Filter housing may be pre-filled with purified water to decrease occurrence of air bubbles. Remove Air From Filter Housing 1. 2. 3. Prepare for Operation 1. 5. Check for air bubbles at the For information on the startup procedure, refer to the ARCHITECT System Operations tubing attached to the degasser Manual, Section 5: Operating instructions, Subsection: System startup, pause, and manifold and the tubing going to shutdown. the probe wash pumps. Check for air bubbles in the tubing on both side of the 12 VDC CKD (DI Water Cuvette Wash) Valve. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. If air bubbles are observed, select the As Needed category from the Maintenance screen under the System icon. Perform 2132 Flush Water Lines. 1. Install the left rear cover. 2. 3. 4. Install Rear Power ON the instrument and wait for the water to fill the filter housing. Loosen the top right bolt (inlet side) [6] on the cap to allow air to escape as the filter housing is filled with water. Note Make sure you remove any remaining air through the bolt. There should not be any bubbles in the filter. Tighten the bolt. [6] Covers Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 G63 - Tubing / Fitting / Filter Verify tubing / fitting / filter is securely installed and does not obstruct instrument components. Ensure no leaks and/or bubbles are in the tubing / fitting / filter. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.04 Pump Head, Water Bath Fill Version - 204738-101_4895_1 List/Part Numbers List/Part Number Description 7-202681-01 Pump Head, Water Bath Fill Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.04 Pump Head, Water Bath Fill Time Required Not Assessed Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Preparation 1. 2. 3. 4. Remove Water Bath Fill Pump Head 1. 2. 3. Reference Drain the water bath For further information, refer to the ARCHITECT System Operations (System/Diagnostics/Modules/1101 Robotics Manual, Section 1: Use or function, Subsection: System status. Test Tool - Option 11). The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the rear left cover. Note Protect the pump from spills. Clamp tubing. Remove the peristaltic pump. Release the clamps to remove the tubing from the peristaltic tubing. Replacement Action Steps Install Water Bath Fill Pump Head 1. 2. 3. Insert the tubing into the peristaltic tubing. Secure the tubing with the clamps. Install the water bath fill pump head. Prepare for Operation 1. Power ON the processing module. Note The system control center power must be ON prior to turning on the Reference For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 2. 3. 4. 5. 6. processing module to ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Select the As Needed category from the Maintenance screen under the System icon. Perform 2134 Change Water Bath. Check for leaks. Install the rear left cover. Verification Procedures Order VP Description VP Detail / Note 1 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.05 Tubing, Pump Head Water Bath Fill Version - 204738-101_4896_2 List/Part Numbers List/Part Number Description 7-202682-01 Tubing, Pump Head Water Bath Fill 7-202682-02 Tubing, Pump Head Water Bath Fill Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.05 Tubing, Pump Head Water Bath Fill Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 5 Min Tools/Materials Standard Tool Kit Absorbent Material WARNING Potential Biohazard WARNING Chemical Hazard WARNING Splash/Spray Hazard Removal Action Steps Reference Preparation 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. 3. 4. VIDEO Remove Water Bath Fill Pump Head and Tubing Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 10). The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the rear left cover. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Water Bath Fill Pump Head and Tubing Note Use absorbent towel to Protect the pump from spills. Clamp tubing. Remove the peristaltic pump. Release the clamps to remove the tubing from the peristaltic tubing. Remove the peristaltic tubing. 1. 2. 3. 4. Replacement Action VIDEO Install Water Bath Fill Tubing and Pump Head Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Water 1. Bath Fill Tubing and Pump 2. Head 3. 4. 5. Prepare for Operation 1. Apply supplied grease to tubing and center hub. Install the peristaltic tubing ensuring is centered on the center hub rollers. Insert the tubing into the peristaltic tubing Secure the tubing with the clamps. Install the water bath fill pump head. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT System Operations Note The system control Manual, Section 5: Operating instructions, Subsection: System startup, pause, and center power must be shutdown. ON prior to turning on the processing module 2. 3. 4. 5. 6. to ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Select the As Needed category from the Maintenance screen under the System icon. Perform 2134 Change Water Bath. Check for leaks. Install the rear left cover. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 G63 - Tubing / Fitting / Filter Verify tubing / fitting / filter is securely installed and does not obstruct instrument components. Ensure no leaks and/or bubbles are in the tubing / fitting / filter. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E3.06 Check Valve, Water Bath Fill Version - 204738-101_4897_1 List/Part Numbers List/Part Number Description 7-203638-01 Check Valve, Water Bath Fill Inspect tools for damage, ensure calibration is not expired and replace if necessary. E3.06 Check Valve, Water Bath Fill Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Material WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Preparation 1. 2. 3. Remove Check Valve 1. 2. 3. Replacement Reference The processing module must be in Stopped or For further information, refer to the ARCHITECT System Operations Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the rear left cover. Clamp the tubing upstream (water bath side, away from the water bath fill peristaltic pump) of the check valve to prevent the water bath from draining. [1] Remove the clamps on each side of the check valve. Remove the check valve. Action Steps Install Check Valve 1. 2. 3. Prepare for Operation 1. 2. 3. Reference Install the check valve into the processing module. Note There is a molded arrow located on the rim of the valve indicating flow direction. The valve must be installed with the tip of the arrow facing towards the water bath and the base of the arrow facing the water bath fill pump. The flow of water is from the peristaltic pump to the water bath. Secure the check valve in place with the clamps. Remove the clamp used to prevent the water bath from draining. Turn on the incoming water source. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be on prior to turning on the processing module to ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E4.01 Pump, Wash Solution Version - 204738-101_4898_1 List/Part Numbers List/Part Number Description 7-206731-01 Pump, Wash Solution 7-93141-02 Pump, Wash Solution (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. E4.01 Pump, Wash Solution Time Required Not Assessed Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Pump, Wash Solution Note 1. 2. 3. 4. 5. 6. 7. 8. Use absorbent material to protect pump against spills. Mark the tubing [1] L (left) and R (right) going to the side ports of the check valve. Clamp the tubing [1] marked L and R going to the side ports of the check valves on the 1mL syringes. Unscrew this tubing from the side ports of the check valves. Unscrew the tubing [2] labeled 2 and 3 from the top of the small syringe unit. Remove the two screws securing the pump to the plate. [3] Slide the pump out until the connectors (DTPM and DTPS) are accessible on the left side of the pump. Note These cables tend to be short and there is not a service loop. Wires in plug DTPS are small and can be damaged easily. Disconnect DTPM and DTPS from the connectors. Remove the pump from the processing module. Replacement Action Steps Install Pump, Wash Solution 1. 2. 3. 4. 5. Slide the pump in the processing module far enough to allow access to the plugs. Connect DTPS and DTPM to the connectors on the connector plate. Install the pump to the frame using two screws. Connect the tubing [2] labeled 2 to the left top of the small syringe and connect the tubing [2] labeled 3 to the right top of the small syringe. Connect the tubing [1] marked L and R to the side ports of the check valves on the 1mL syringes and remove the clamps. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be on prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close supply center (left front) door. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E4.02 Syringe Assy, Wash Solution Version - 204738-101_4899_1 List/Part Numbers List/Part Number Description 7-93143-01 Syringe Assy, wash solution Inspect tools for damage, ensure calibration is not expired and replace if necessary. E4.02 Syringe Assy, Wash Solution Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open supply center (left front) door. 2. Remove Small Syringe Unit Note 1. 2. 3. Use absorbent material to protect the wash solution pump from spills. Disconnect tubing [1] labeled 2 and 3 from the top and the front of the small syringe unit. Loosen the thumbscrew [2] securing the small syringe unit to the wash solution pump. While removing the small syringe unit, pull the unit forward and lift. Replacement Action Steps Install Small Syringe Unit 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. Reference While installing the small syringe unit use a down and to the rear movement to engage syringe plungers into drive mechanism. Install the small syringe unit to the wash solution pump using the thumbscrew. [2] Connect the tubing[1] to the front of the small syringe. Connect the tubing [1] to the top of the small syringe. Note Tubing labeled 2 is to the left and tubing labeled 3 is to the right. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from Stopped to System Operations Manual, Section 5: Operating instructions, Ready, perform the Startup procedure. Subsection: System startup, pause, and shutdown. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close the supply center (left front) door. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E4.03 Seal, Water Line Syringe (Wash Solution Pump) Version - 204738-101_4900_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E4.03 Seal, Water Line Syringe (Wash Solution Pump) Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Open supply center (left front) door. 2. Remove Small Syringe Unit Note 3. Use absorbent material to protect the wash solution pump from spills. Disconnect tubing [1] labeled 2 and 3 from the top and the front of the small syringe unit. Loosen the thumbscrew [2] securing the small syringe unit to the wash solution pump. Remove the small syringe unit. [3] 1. Remove seal [1] from syringe. 1. 2. Remove Seal Replacement Action Steps Install Seal 1. Install seal [1] in syringe. Install Small Syringe Unit 1. 2. Replace the small syringe unit. Secure the small syringe unit [3] to the wash solution pump using the thumbscrew. [2] Connect the tubing[1] to the front of the small syringe. Connect the tubing [1] to the top of the small syringe. Note Tubing labeled 2 is to the left and tubing labeled 3 is to the right. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. Reference To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from Stopped to System Operations Manual, Section 5: Operating instructions, Ready, perform the Startup procedure. Subsection: System startup, pause, and shutdown. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close the supply center (left front) door. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E5.01 Pump, ICT Reference Version - 204738-101_4901_1 List/Part Numbers List/Part Number Description 7-206732-01 Pump, ICT Ref Solution Inspect tools for damage, ensure calibration is not expired and replace if necessary. E5.01 Pump, ICT Reference Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Pump, ICT Reference Note 1. 2. 3. 4. 5. Use absorbent material to protect pump against spills. Unscrew the check valves [1] from the top of the 1 mL syringes (tubing labeled 13-left, 12-right). Remove the two (2) screws [2] securing the pump to the plate. Carefully slide the pump [3] out of the processing module until the connectors (ISEMM and ISEMS) are accessible. Note These cables (located on the left side) tend to be short and there is not a service loop. Wires to plug ISEMS are small and can be damaged easily. Disconnect the ISEMM and ISEMS cables. Remove the pump [3] from the processing module. Replacement Action Steps Install Pump, ICT Reference 1. 2. 3. 4. Prepare for Operation 1. Reference Slide the pump [3] in the processing module far enough to allow access to the plugs. Connect the plugs ISEMM and ISEMS to the connectors on the connector plate. Install the pump to the frame using two (2) screws . [2] Connect the check valves [1] (tubing labeled 13-left, 12-right) to top of the 1 mL syringes. Power ON the processing module. Note The system control center power For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, 2. 3. 4. 5. 6. must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2131 Flush ICT Cup. Check for leaks. Close the supply center door. Verification Procedures Order VP Description VP Detail / Note 1 2131 - Flush ICT Cup Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E6.01 Pump, ICT Aspiration w/Pinch Valve (RoHS) Version - 204738-101_4902_1 List/Part Numbers List/Part Number Description 7-206733-01 Pump, ICT Aspiration w Pinch Valve 7-93145-03 Pump, ICT Aspiration w Pinch Valve (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. E6.01 Pump, ICT Aspiration w/Pinch Valve (RoHS) Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Pump, ICT Aspiration Note 1. 2. 3. 4. Remove Pump, ICT Aspiration (continued) 1. 2. Use absorbent material to protect pump against spills. Unscrew the check valve [1] from the top of the 1 mL syringe, on the left side of the aspiration pump. Unlock the tubing clamp [2] and remove the tubing from the two fittings [3]. Remove the two screws securing the pump to the plate. Carefully slide the pump out of the processing module until the connectors are accessible. Disconnect the ISESM, ISESS, and ISPV cables. Remove the pump from the processing module. Replacement Action Steps Install Pump, ICT Aspiration 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. 5. Reference Slide the pump in the processing module far enough to allow access to the plugs. Connect the plugs ISESM, ISESS, and ISPV cables. Install the aspiration pump to the frame using two screws. Route tubing through the tubing clamp [2] and attach the tubing to the fittings [3]. The tubing labeled "NO" goes to the inside and the tubing labeled "NC" goes to the outside. Note Trim the tip of the tubing by approximately 5mm as required to ensure a good fit between the tubing and the fitting. Note Fluid flows from the ICT module into the NC port and out from the NO port to the waste drain. Lock tubing clamp [2]. Connect the check valve with tubing 14 to the the 1 mL syringe on the left. [1] Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Maintenance, As Needed, 6063 Flush ICT Module. Check for leaks. Close the supply center door. Verification Procedures Order VP Description VP Detail / Note 1 2131 - Flush ICT Cup Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in Expected Results specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E6.02 Pump Pinch Valve, ICT Aspiration Version - 204738-101_4903_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E6.02 Pump Pinch Valve, ICT Aspiration Time Required 30 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Pump, ICT Aspiration Note 1. 2. 3. 4. 5. 6. Use absorbent material to protect pump against spills. Unscrew the check valve [1] from the top of the 1 mL syringe, on the left side of the aspiration pump. Unlock the tubing clamp [2] and remove the tubing from the two fittings [3]. Remove the two screws securing the pump to the plate. Carefully slide the pump out of the processing module until the connectors are accessible. Disconnect the ISESM, ISESS, and ISPV cables. Remove the pump from the processing module. Replacement Action Steps Remove the Pinch Tubing 1. 2. 3. 4. Remove Pinch Valve 1. 2. Install Pinch Valve 1. 2. Install Pump, ICT Aspiration 1. 2. 3. 4. Loosen the two screws securing the guide plate [2] and rotate the guide plate away from the pinch tubing. Pull on the tubing [3] to disconnect the tubing from the two fittings. Note Do not remove the fittings from the locking plate. Slide the tubing out from the normally opened (NO) port of the valve. Push on the button [4] to un-pinch the tubing on the normally closed (NC) port of the valve and remove the tubing. Compress locking tabs on the connector and remove the ISEV connector from the connector plate. Remove two (2) outer screws securing the valve to the pump frame and remove the valve. Install the valve in the hole for the valve on the ICT aspiration pump assembly and secure using two (2) screws. Insert the connector ISEV in the connector plate. Slide the pump in the processing module far enough to allow access to the plugs. Connect the ISESM, ISESS and ISPV cables. Install the aspiration pump to the frame using two (2) screws. Route tubing through the tubing clamp [2] and attach the tubing to the fittings [3]. The tubing labeled with "NO" goes to the inside Reference 5. 6. Install Pinch Tubing 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. and the tubing labeled "NC" goes to the outside. Note Ensure a good fit between the tubing and the fitting. Note Fluid flows from the ICT module into the NC port and out from the NO port to the waste drain. Lock tubing clamp [2]. Connect the check valve with tubing 14 to the the 1 mL syringe on the left. [1] Slide the pinch tubing back into the NO port of the valve. Push on the button and slide the other tubing into the NC port of the valve. Connect the tubing onto the fittings. Rotate the guide plate until it rests against the pinch valve and tighten the two (2) screws. Power ON the processing module. For further information, refer to the ARCHITECT System Operations Note The system control center power Manual, Section 1: Use or function, Subsection: System status. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6063 Flush ICT Module. Check for leaks. Close the left front door. Verification Procedures Order VP Description VP Detail / Note 1 6063 - Flush ICT Module Check for leaks. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E6.03 ICT Pinch Valve Tubing Version - 204738-101_4904_2 List/Part Numbers List/Part Number Description 7-204068-01 Tubing, ICT Pinch Valve Inspect tools for damage, ensure calibration is not expired and replace if necessary. E6.03 ICT Pinch Valve Tubing Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 10 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Splash/Spray Hazard Removal Action Steps Reference Prerequisite The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. 2. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door to gain access to the ICT Aspiration Pump. VIDEO Remove Note the Pinch Tubing Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove the Pinch Tubing 1. Disconnect the tubing [1] from the top of the 1 mL syringe, on the right side of the aspiration pump. Loosen the two screws securing the guide plate [2] and rotate the guide plate away from the pinch tubing. Pull on the tubing [3] to disconnect the tubing from the two fittings. Note Do not remove the fittings from the locking plate. Slide the tubing out from the normally opened (NO) port of the valve. Push on the button to unpinch the tubing on the normally closed (NC) port of the valve and remove the tubing. 2. 3. 4. 5. Replacement Action Steps VIDEO Install Pinch Tubing Note Reference Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Pinch Tubing 1. 2. 3. 4. 5. Connect the tubing to the 1 mL syringe and ensure that the tubing is not twisted. Note The tubing will not be twisted if it is installed perpendicular (at 90 degrees) to the tabs on the 1 mL syringe barrel. Slide the pinch tubing into the NO port of the valve. Push on the button [2] and slide the other tubing into the NC port of the valve. Push the pinch tubing onto the two fittings. Rotate the guide plate until it rests against the pinch valve and tighten the two screws. Prepare for Operation 1. 2. 3. 4. 5. 6. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Maintenance, As Needed, 6063 Flush ICT Module. Check for leaks. Close the supply center (left front) door. Log on as Operator. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6063 - Flush ICT Module Check for leaks. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E7.01 Pump, Probe Wash, Bellows Type Version - 204738-101_4905_1 List/Part Numbers List/Part Number Description 2-89142-02 Cover,Bellows pump probe wash Inspect tools for damage, ensure calibration is not expired and replace if necessary. E7.01 Pump, Probe Wash, Bellows Type Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or For further information, refer to the ARCHITECT System Operations Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Inner Probe Wash Pump Reference Note 1. 2. 3. 4. 5. 6. Use absorbent material to protect pump from spills. Clamp the inlet and outlet tubing [1]. Disconnect the inlet and outlet tubing by unscrewing the two elbow fittings [2]. Remove the two (2) screws [3] securing the pump to the processing module base. Pull the pump forward until the connectors are accessible. Disconnect the two cables at connectors PRWPM and PRWPS. Remove the pump from the processing module. Replacement Action Steps Install Inner Probe Wash Pump 1. 2. 3. Connect the cables for connectors PRWPM and PRWPS. Install the pump in the processing module and secure using the two (2) screws [3]. Connect the inlet and outlet tubing to the pump by reconnecting the two elbow fittings [2]. Note Ensure poppet valves and o-rings are properly oriented. The upper poppet points toward the bellows; the lower poppet points away from the bellows; the o-ring is placed Reference 4. Prepare for Operation 1. 2. 3. between the poppet and the elbow fittings. Remove clamps from the inlet and outlet tubing [1]. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Maintenance, As Needed, 6053 Probe Water Wash. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E7.02 Complete Bellows Set, Probe Wash Pump Version - 204738-101_4906_2 List/Part Numbers List/Part Number Description 2-89055-02 Bellows set,incl rod & fitting Inspect tools for damage, ensure calibration is not expired and replace if necessary. E7.02 Complete Bellows Set, Probe Wash Pump Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Splash/Spray Hazard Removal Action Steps Reference Prerequisite The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. 2. VIDEO Remove Bellows Pump Cover Tubing and Bellows Set Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. Note Video contains no audio sound. Note Do Not remove the pump. The pump shown is removed from the instrument for demonstration purposes only. (If the video does not display, or to view the video full size: Click Here) Remove Note Bellows Pump Cover and Tubing Use absorbent material to protect the pump from spills. 1. 2. 3. Remove the two (2) screws from the front of the cover [1] and remove the cover from the bellows pump. Clamp the inlet and outlet tubing. [2] Disconnect the inlet and outlet tubing by unscrewing the two elbow fittings. [3] Remove Bellows Set Note Use absorbent material to protect the pump from spills. 1. 2. 3. 4. Remove the four (4) screws securing the gear guard [3] and remove guard. Remove the screw [4] from the end of the bellows rod and remove the insert and the thrust washer [5]. Remove the two (2) screws [6] from the top of the bellows set [7] and remove the bellows set from the pump. Remove the Poppet valves and O-rings. Note Poppet valves and O-rings may be removed and reinstalled into a new bellows set or replaced at the discretion of the FSR. Replacement Action Steps VIDEO Note Install Bellows Set, Tubing and Pump Cover Reference Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Bellows Set 1. 2. 3. Install the thrust washer and the insert [5] in the end of the bellows rod. Install the bellows set [7] in the pump by placing the rod in the mounting position and securing the top of the bellows set to the top of pump using two (2) screws [6]. Secure the thrust washer and insert it using one (1) screw. [4] Note Water may be added to the bellows to prevent the occurrence of bubbles. 4. 5. Install Bellows Pump Tubing and Cover 1. 2. 3. Prepare for Operation 1. 2. 3. 4. Reinstall the poppet valves and O-rings previously removed, or install new sets. Install the gear guard [3] to the pump using four (4) screws. Connect the inlet and outlet tubing to the pump by reconnecting the elbow fittings. Remove the clamps from the inlet and outlet tubing. Install the cover on the bellows pumps and secure it with two (2) screws. Perform the Startup procedure For information on the startup procedure, refer to the ARCHITECT System Operations to change the status of the Manual, Section 5: Operating instructions, Subsection: System startup, pause, and processing module and sample shutdown. handler from Stopped to Ready. Select System, Maintenance, As Needed, 6053 Probe Water Wash. Check for leaks. Close the supply center (left front) door. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015• ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E7.03 Bellows, Probe Wash Pump Version - 204738-101_4907_2 List/Part Numbers List/Part Number Description 7-204102-01 Pump, Probe Wash, Bellows Type (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. E7.03 Bellows, Probe Wash Pump Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Splash/Spray Hazard Removal Action Steps Prerequisite The processing module must be in For further information, refer to the ARCHITECT System Operations Manual, Section Stopped or Offline status. 1: Use or function, Subsection: System status. Preparation 1. 2. VIDEO Remove Pump Cover and Bellows Reference Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Video Note Do Not Remove Pump Assembly. Pump shown removed from instrument for ease of demonstrating bellows removal and replacement steps. Remove Pump Cover and Bellows Note Use absorbent material to protect the pump from spills. 1. Clamp the inlet and outlet tubing. Remove the two (2) screws from the front of the cover and remove cover. Unscrew the retaining ring [1] from the bottom of the bellows. Remove the two (2) screws [2] from the top of the bellows assembly. [4] Remove the bellows assembly and unscrew the bellows [3] from the top of the bellows assembly. [4] Remove the O-ring [5] from inside the top of the bellows assembly. [4] 2. 3. 4. 5. 6. Replacement Action VIDEO Install Pump Cover and Bellows Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Pump Cover and Bellows 1. Install the O-ring [5] inside the top of the bellows assembly. Note Ensure that the cap [6] is inserted into the bellows prior to screwing the bellows in the top of the bellows assembly. 2. Screw the bellows [3] in the top of the bellows assembly. [4] 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. Install the bellows assembly to the pump using two (2) screws. [2] Attach the bellows to the rod with the retaining ring. [1] Install the cover on the bellows pumps and secure using two (2) screws. Remove the clamps from the inlet and outlet tubing. Perform the Startup procedure For information on the startup procedure, refer to the ARCHITECT System Operations to change the status of the Manual, Section 5: Operating instructions, Subsection: System startup, pause, and processing module and sample shutdown. handler from Stopped to Ready. Select System, Maintenance, As Needed, 6053 Probe Water Wash. Check for leaks. Close the supply center (left front) door. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E7.04 L-bow Fitting, Probe Wash Pump Version - 204738-101_4908_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E7.04 L-bow Fitting, Probe Wash Pump Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or For further information, refer to the ARCHITECT System Operations Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove L-Bow Fitting Reference Note 1. 2. 3. Use absorbent material to protect pump from spills. Clamp the inlet and outlet tubing [1]. Remove the tubing [1] from the elbow [2]. Unscrew the elbow fittings from the bellows [3]. Replacement Action Steps Install L-Bow Fitting 1. 2. 3. Screw the elbow fittings to the bellows [3]. Note Ensure poppet valves and o-rings are properly oriented. The upper poppet points toward the bellows; the lower poppet points away from the bellows; the o-ring is placed between the poppet and the elbow fittings. Attach the tubing [1] to the elbow [2]. Remove clamps from the inlet and outlet tubing [1]. Reference Prepare for Operation 1. 2. 3. 4. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Select System, Maintenance, As Needed, 6053 Probe Water Wash. Check for leaks. Close the supply center (left front) door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E8.01 Pump, Peristaltic, High Conc Waste Version - 204738-101_4909_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E8.01 Pump, Peristaltic, High Conc Waste Time Required 30 Min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove Peristaltic H.C. Waste Pump Head Note 1. 2. Remove Deck Covers 1. 2. Use absorbent material to protect the pump from spills. Turn each peristaltic pump head counterclockwise (CCW) about 1/16 of a turn. Pull upward on the pump head to remove the head. Remove Front Right Cover [C]. Remove Rear Right Cover [K]. Remove Peristaltic Pump Connectors 1. 2. Remove Peristaltic Pump Assembly 1. 2. 3. Disconnect plug [1] HCWPM from the connector. Remove connector HCWPM [2] from the mounting bracket. Remove the four (4) screws [1] securing the pump assembly. Disconnect sensor plug HPPITCH [2] from the pump assembly. Remove the pump assembly. Replacement Action Steps Install Peristaltic Pump Assembly 1. 2. Install Peristaltic Pump Connectors 1. 2. Connect sensor plug HPPITCH [2] to the pump assembly. Install the peristaltic pump assembly to the mounting bracket using four (4) screws. [1] Install connector HCWPM [2] to the mounting bracket. Connect plug HCWPM [1] to the connector. Reference Install Deck Covers 1. 2. Install Front Right Cover [C]. Install Rear Right Cover [K]. Install Peristaltic H.C. Waste Pump Head 1. Align the notch of the roller assembly with the peristaltic motor tang. Align the attachment locking tabs on the peristaltic pump head with the recesses on the pump. Turn head clockwise (CW) about 1/16 of a turn until it snap-locks in place. Attach tubing from the Cuvette Wash Station HCW nozzle and the tubing going to the HCW drain/container. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Select System, Maintenance, As Needed, 6052 Wash Cuvettes Test. Check for leaks. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks Expected Results N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E8.02 Pump, Peristaltic Head Version - 204738-101_4910_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E8.02 Pump, Peristaltic Head Time Required 30 Min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove Peristaltic H.C. Waste Pump Head Note 1. 2. 3. 4. 5. Use absorbent material to protect the pump from spills. Clamp tubing. Remove any hose bands. Remove the tubing from the nipples of the peristaltic pump head. Turn each peristaltic pump head. counterclockwise (CCW) about 1/16 of a turn. Pull upward on the pump head to remove the head. Replacement Action Steps Install Peristaltic H.C. Waste Pump Head 1. 2. 3. 4. Align the notch of the roller assembly with the peristaltic motor tang. Align the attachment locking tabs on the peristaltic pump head with the recesses on the pump. Turn head clockwise (CW) about 1/16 of a turn until it snap-locks in place. Attach tubing. Secure with hose bands (if removed) and remove clamps. Reference Prepare for Operation 1. 2. 3. 4. 5. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes Test. Check for leaks. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E8.03 Tubing, Peristaltic Head Version - 204738-101_4911_2 List/Part Numbers List/Part Number Description 7-202464-01 Tubing, Peristaltic Head Inspect tools for damage, ensure calibration is not expired and replace if necessary. E8.03 Tubing, Peristaltic Head Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 15 Min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Chemical Hazard WARNING Splash/Spray Hazard Removal Action Steps Prerequisite The processing module status For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. must be Ready, Stopped or Offline. Preparation Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) VIDEO Remove Pump Head and Tubing Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Pump Head and Tubing Note 1. 2. 3. 4. 5. Use absorbent material to protect the pump from spills. Remove any hose band. Remove the tubing from the fittings of the peristaltic pump head. Turn the peristaltic pump head counterclockwise (CCW) about 1/16 of a turn. Pull upward on the pump to remove the head. Pull out the plastic housing with the rollers attached and discard the tubing with fittings. Replacement Action VIDEO Install Pump Head Tubing with Fittings and Peristaltic Pump Head Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Peristaltic Pump Head Tubing with Fittings 1. 2. 3. 4. 5. Apply supplied lubricant to tubing and center shaft of pump housing. Insert upper fitting of tubing into pump housing while aligning the flat cut surface of the upper fitting with the inlet opening of the pump housing. Install roller assembly onto the center shaft of the pump housing. Note The cable tie should fit into the recess of the pump housing. Pull up the roller assembly and push in the tubing with the thumb while turning the roller assembly clockwise (CW). Insert lower joint of tubing into the pump housing while aligning the flat cut surface of the lower joint with the outlet opening of the pump housing. Note The cable tie should fit into the recess of Note Install Peristaltic Pump Head 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. the pump housing. Once tubing is installed, ensure tubing is not twisted. Twisted tubing in the pump housing can cause premature failure of the tubing. Align the notch of the roller assembly with the peristaltic motor tang. Align the attachment locking tabs on the peristaltic pump head with the recesses on the pump. Turn head clockwise (CW) about 1/16 of a turn until it snap-locks in place. Attach tubing from the Cuvette Wash Station HCW nozzle and the tubing going to the HCW drain/container. Perform the Startup procedure For information on the startup procedure, refer to the ARCHITECT System Operations to change the status of the Manual, Section 5: Operating instructions, Subsection: System startup, pause, and processing module and sample shutdown. handler from Stopped to Ready. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E9.01 Pump, Cuvette Wash, Bellows Version - 204738-101_4912_1 List/Part Numbers List/Part Number Description 7-93146-02 Pump, Cuvette Wash, Bellows (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. E9.01 Pump, Cuvette Wash, Bellows Time Required 20 Min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Cuvette Wash Pump Note 1. 2. 3. 4. 5. 6. Use absorbent material to protect the pump from spills. Clamp the inlet and outlet tubing. [1] Disconnect the inlet and outlet tubing by unscrewing the elbow fittings. [2] Remove the two screws [3] securing the pump [4] to the processing module base. Pull the pump forward until the connectors are accessible. Disconnect the cables WSPM and WSPS. Remove the pump from the processing module. Replacement Action Steps Install Cuvette Wash Pump 1. 2. Connect the cables WSPM and WSPS. Install the pump in the processing module and secure using two screws. Connect Cuvette Wash Pump Tubing and Remove Clamps 1. Connect the inlet and outlet tubing [1] to the pump by reconnecting the elbow fittings [2]. Note Ensure poppet valves and o-rings are properly oriented. The upper Reference 2. Prepare for Operation 1. 2. poppet points toward the bellows; the lower poppet points away from the bellows; the o-ring is placed between the poppet and the elbow fittings. Remove clamps from the inlet and outlet tubing [1]. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E9.02 Complete Bellows Set, Cuvette Wash Pump Version - 204738-101_4913_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E9.02 Complete Bellows Set, Cuvette Wash Pump Time Required 20 Min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Bellows Pump Cover 1. 2. Remove Inlet/Outlet Tubing Remove the two screws from the front of the cover [1]. Remove the cover from the bellows pump. Note 1. 2. Remove Bellows Set Use absorbent material to protect pump from spills. Clamp the inlet and outlet tubing [1]. Disconnect the inlet and outlet tubing by unscrewing the two elbow fittings [2]. Note 1. 2. 3. Use absorbent material to protect pump from spills. Remove the four screws securing the gear guard [3] and remove guard. Remove the thrust washer [2]. Remove the two screws [3] from the top of the bellows set [4], and remove the bellows set from the pump. Replacement Action Steps Install Bellows Set 1. 2. 3. 4. 5. Reference Install the bellows set [4] in the pump by placing the rod on motor shaft and securing the top of the bellows set to the top of pump using two screws [3]. Install the thrust washer [2] on the shaft. Install the boss, and tighten the two set screws. Connect the inlet and outlet tubing to the pump by reconnecting the elbow fittings. Remove the clamps from the inlet and outlet tubing. Install Bellows Pump Cover 1. Install the cover on the bellows pumps and secure using two screws. Prepare for Operation 1. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from Stopped System Operations Manual, Section 5: Operating instructions, to Ready, perform the Startup procedure. Subsection: System startup, pause, and shutdown. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close the left front door. 2. 3. 4. 5. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per Expected Results released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. E9.03 Bellows, Cuvette Wash Pump Version - 204738-101_4914_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. E9.03 Bellows, Cuvette Wash Pump Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 18 Min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Chemical Hazard WARNING Splash/Spray Hazard Removal Action Steps Reference Prerequisite The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. 2. VIDEO Remove Pump Cover and Bellows Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. Note Do Not Remove Pump Assembly. Pump shown removed from instrument for ease of demonstrating bellows removal and replacement steps. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Pump Note Use absorbent Cover and Bellows 1. 2. 3. 4. 5. 6. material to protect pump from spills. Clamp the inlet and outlet tubing Remove the two (2) screws from the front of the cover and remove cover. Unscrew the retaining ring [1] from the bottom of the bellows. Remove the two screws [2] from the top of the bellows. Remove the bellows assembly and unscrew the bellows [4] from the top of the bellows assembly.[3] Remove the O-ring [5] from the inside of the bellows. Replacement Action VIDEO Install Pump Cover and Bellows Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Pump Cover and Bellows 1. Install the O-ring [5] inside the top of the bellows. Note Ensure that the cap [6] is inserted into the bellows prior to screwing the bellows in the top of the bellows assembly. 2. 3. 4. 5. Screw the bellows [4] in the top of the bellows assembly. [3] Install the bellows assembly to the pump using two screws. [2] Attach the bellows to the rod with the retaining ring. [1] Install the cover on the bellows pumps and secure using two (2) screws. Prepare for Operation 6. Remove the clamps from the inlet and outlet tubing. 1. To change the status of the For information on the startup procedure, refer to the ARCHITECT System Operations processing module and sample Manual, Section 5: Operating instructions, Subsection: System startup, pause, and handler from Stopped to shutdown. Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close the left front door. 2. 3. 4. 5. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. F1.01 Acid, Alkaline Valves Version - 204738-101_4915_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. F1.01 Acid, Alkaline Valves Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard WARNING Chemical Hazard WARNING Splash/Spray Hazard Removal Action Steps Prerequisite The processing module must be in For further information, refer to the ARCHITECT System Operations Manual, Section Stopped or Offline status. 1: Use or function, Subsection: System status. Note This procedure is for the four valves in the bulk solution assembly. Preparation 1. 2. 3. 4. 5. VIDEO Remove Wash Solution Valve Reference Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. Remove the tubing from the ICT reference bottle, alkaline wash bottle, and acid wash bottle. Wrap each tube with an absorbent towel. Remove the ICT reference bottle, alkaline wash bottle, and acid wash bottle from the weight sensor platforms. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Wash Solution Valve 1. Remove the tubing from the appropriate valve (the valve being removed). Unplug the appropriate plug [1] (ALK1, ACD2, ALK2, or ACD1) for the valve being removed and remove cables from clamp. Loosen the two (2) screws, [2] underneath the port labels securing the connector plate to the frame, lift up on the connector plate and bring forward. Remove the left screw from the mounting plate for the valve being removed and loosen the right screw. Slide the valve down and remove from the mounting plate. 2. 3. 4. 5. Replacement Action Steps VIDEO Install Note Wash Solution Valve Reference Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Wash 1. Solution Valve 2. 3. 4. 5. Slide the valve in position on the mounting plate and secure using two (2) screws (right and left). Install the plate to the frame using two (2) screws. [2] Connect the appropriate connector [1] (ALK1, ACD2, ALK2, or ACD1). Secure cables in clamp. Install the tubing for the appropriate valve. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. 8. 9. Prepare for Operation 1. Install the ICT reference bottle, alkaline wash bottle, and acid wash bottle on the weight sensor platforms. Install the tubing in the ICT reference bottle, alkaline wash bottle, and acid wash bottle. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2155 Flush Bulk Solutions. Perform 6052 Wash Cuvettes. Check for leaks. Close the supply center (left front) door. Perform the Startup procedure For information on the startup procedure, refer to the ARCHITECT System Operations to change the status of the Manual, Section 5: Operating instructions, Subsection: System startup, pause, and processing module and sample shutdown. handler from Stopped to Ready. Verification Procedures Order VP Description 1 2155 - Flush Bulk Solutions 2 6052 - Wash Cuvettes N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Check for leaks Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. F1.02 Weight Sensor Version - 204738-101_4916_1 List/Part Numbers List/Part Number Description 7-93485-01 Weight sensor, N/S, C8 Inspect tools for damage, ensure calibration is not expired and replace if necessary. F1.02 Weight Sensor Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. Remove the tubing from the ICT reference bottle, alkaline wash bottle, and acid wash bottle. Wrap each tube with an absorbent towel. Remove the ICT reference bottle, alkaline wash bottle, and acid wash bottle from the weight sensor platforms. 2. 3. 4. 5. Remove Weight Sensor Platform 1. 2. 3. 4. Remove the two (2) screws [1] securing the base to the processing module frame. Slide the platform [2] forward and invert to gain access to the sensor plugs [3]. Undo cable ties [4]. Disconnect the plugs [3] from the sensors: • ALK, (alkaline) • ACD, (acid) • MCC, (ICT reference) Remove the weight sensor platform [2]. Replacement Action Steps Install Weight Sensor Platform 1. 2. 3. Prepare for Operation 1. 2. 3. Reference Connect the plugs [3] to the sensors • ALK, (alkaline) • ACD, (acid) • MCC,(ICT reference) Reconnect cable ties [4]. Install the base to the frame using two (2) screws. [1] Install the ICT reference bottle, alkaline For information on the startup procedure, refer to the ARCHITECT wash bottle, and acid wash bottle on the System Operations Manual, Section 5: Operating instructions, weight sensor platforms. Subsection: System startup, pause, and shutdown. Install tubing in the ICT reference bottle, alkaline wash bottle, and acid wash bottle. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Verification Procedures Order VP Description 2 2155 - Flush Bulk Solutions 1 P-612 Bulk Solution Weight Sensor Adjustment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G1.01 Drive, Sample Syringe Version - 204738-101_4917_1 List/Part Numbers List/Part Number Description 7-93154-01 Drive, Sample Syringe Inspect tools for damage, ensure calibration is not expired and replace if necessary. G1.01 Drive, Sample Syringe Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Sample Syringe Drive Note 1. 2. 3. 4. 5. Use absorbent material to protect sample syringe drive against spills. Remove the plastic cover [4]. Unscrew the tubing [1] from the top and side of the sample syringe. Remove the two (2) screws [2] at the bottom of the assembly. Slide syringe drive assembly out of the processing module for access to the cables and disconnect plugs SP-M and SP-S from syringe drive assembly. [3] Remove the syringe drive assembly from the processing module. Remove Sample Syringe 1. 2. Remove the two (2) screws securing the sample syringe plunger clamp and remove the clamp. [2] Remove the screw securing the syringe and remove the syringe. [3] Replacement Action Steps Install Sample Syringe 1. 2. Install the syringe [3] using one (1) screw. Install the plunger clamp [2] using two screws. Install Sample Syringe Drive 1. Connect plugs SP-M and SP-S to the syringe drive assembly. Slide the syringe drive assembly [3] in the processing module and secure using the two (2) screws [2] at the base of the assembly. Connect the tubing [1] from the probe to the 2. 3. Reference 4. Prepare for Operation 1. 2. 3. 4. 5. 6. top of the sample syringe and from the valve to the side of the sample syringe. Install the plastic cover using two screws. [4] Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G1.02 Spring, Sample Syringe Version - 204738-101_4918_1 List/Part Numbers List/Part Number Description 2-89381-02 Spring,Sample Syringe Shaft Inspect tools for damage, ensure calibration is not expired and replace if necessary. G1.02 Spring, Sample Syringe Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Sample Syringe Note 1. 2. 3. 4. Remove Sample Syringe Spring 1. 2. 3. 4. 5. Use absorbent material to protect sample syringe drive against spills. Unscrew the tubing [1] from the top and side of the sample syringe. Remove the two (2) screws securing the plastic cover and remove the cover. Remove the two (2) screws securing the sample syringe plunger clamp and remove the clamp. [2] Remove the one (1) screw securing the syringe and remove the syringe. [3] Unscrew the nut [5] on the bottom of the syringe block and slide the syringe plunger [6] and seal tips [2, 3] out of the syringe. Remove the o-ring [1] from seal tip #2. Remove seal tip #2 [2], the spacer [7] and seal tip #1 [3] from the plunger. Pull the nut [5] down on the plunger to reveal the spring. [4] Remove the spring from the plunger. [6] Replacement Action Steps Install Sample Syringe Spring 1. 2. 3. 4. Install Sample Syringe 1. 2. 3. 4. Prepare for Operation 1. Reference Install the spring [4] on the plunger. [6] Install seal tip #1 [3], the spacer and seal tip #2 [2] on the plunger. Install the o-ring [1] on seal tip #2. Insert the plunger using the seals in the syringe block and secure using the nut [5] on the bottom. Note The nut must fit flush against the syringe block. Install the syringe [3] using one (1) screw. Install the plunger clamp [2] using two screws. Install the plastic cover using two screws. Connect the tubing [1] from the probe to the top of the sample syringe and from the valve to the side of the sample syringe. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 2. 3. 4. 5. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G1.03 Seal, Water Line, Syringe (Sample) Version - 204738-101_4919_2 List/Part Numbers List/Part Number Description 2-89347-02 Seal, water line syringe Inspect tools for damage, ensure calibration is not expired and replace if necessary. G1.03 Seal, Water Line, Syringe (Sample) Time Required 18 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Sample Syringe Note 1. 2. 3. 4. Use absorbent material to protect sample syringe drive against spills. Unscrew the tubing [1] from the top and side of the sample syringe. Remove the two (2) screws securing the plastic cover and remove the cover. Remove the two (2) screws securing the sample syringe plunger clamp and remove the clamp. [2] Remove the one (1) screw securing the syringe and remove the syringe. [3] (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Remove Seals 1. Remove the seals from the sample syringe. Replacement Action Steps Install Seals 1. Install the seals in the sample syringe. Install Sample 1. Install the syringe [3] using Reference Syringe 2. 3. 4. one (1) screw. Install the plunger clamp [2] using two screws. Install the plastic cover using two screws. Connect the tubing [1] from the probe to the top of the sample syringe and from the valve to the side of the sample syringe. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Prepare for Operation 1. 2. 3. 4. 5. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G1.04 Syringe, Sample Version - 204738-101_4920_1 List/Part Numbers List/Part Number Description 2-89399-02 Syringe, sample, complete Inspect tools for damage, ensure calibration is not expired and replace if necessary. G1.04 Syringe, Sample Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Sample Syringe Note 1. 2. 3. 4. Use absorbent material to protect sample syringe drive against spills. Unscrew the tubing [1] from the top and side of the sample syringe. Remove the two (2) screws securing the plastic cover and remove the cover. Remove the two (2) screws securing the sample syringe plunger clamp and remove the clamp. [2] Remove the one (1) screw securing the syringe and remove the syringe. [3] Replacement Action Steps Install Sample Syringe 1. 2. 3. 4. Install the syringe [3] using one (1) screw. Install the plunger clamp [2] using two screws. Install the plastic cover using two screws. Connect the tubing [1] from the probe to the top of the sample syringe and from the valve to the side of the sample syringe. Reference Prepare for Operation 1. 2. 3. 4. 5. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G2.01 Drive, Reagent Syringe Version - 204738-101_4921_1 List/Part Numbers List/Part Number Description 7-93156-01 Drive, Reagent Syringe Inspect tools for damage, ensure calibration is not expired and replace if necessary. G2.01 Drive, Reagent Syringe Note This procedure is for both R1 and R2 reagent syringe drives. Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Reagent Syringe Drive Note 1. 2. 3. 4. 5. Remove Reagent Syringe Drive Syringe 1. 2. Use absorbent material to protect reagent syringe drive against spills. Remove the two (2) screws securing the plastic cover. Unscrew the tubing [1] from the top and side of the reagent syringe. Remove the two (2) screws [2] at the bottom of the assembly. Slide syringe drive assembly out of the processing module for access to the following cables. • Disconnect plugs R1 P-M and R1 P-S from R1 drive assembly. • Disconnect plugs R2 P-M and R2 P-S from R2 drive assembly. Remove the syringe drive assembly [3] from the processing module. Remove the two screws securing the reagent syringe plunger clamp and remove the clamp. [2] Remove the screw [4] securing the syringe and remove the syringe. [3] Replacement Action Steps Install Reagent Syringe Drive Syringe 1. 2. Install the syringe [3] using one screw. [4] Install the plunger clamp [2] using two (2) screws. Install Reagent Syringe Drive 1. Connect plugs R1P-M/R2P-M and R1PS/R2P-S to the syringe drive assembly. [3] Slide the syringe drive assembly in the processing module and secure using the two screws [2] at the base of the assembly. Connect the tubing [1] from the probe to the top of the reagent syringe and from the valve to the side of the reagent syringe. Install the plastic cover using two (2) screws. 2. 3. 4. Prepare for Operation 1. Reference Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure 2. 3. 4. 5. 6. proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G2.02 Seal, Water Line, Syringe (Reagent) Version - 204738-101_4922_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. G2.02 Seal, Water Line, Syringe (Reagent) Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Open the left front door. VIDEO Remove Reagent Syringe and Seals Note Video shows steps for the removal of the Sample Syringe and Seals. However, steps are also applicable to the removal of the Reagent Syringe and Seals. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Reagent Syringe and Seals Note 1. 2. 3. 4. Use absorbent material to protect reagent syringe drive against spills. Unscrew the tubing [1] from the side and top of the reagent syringe. Remove the two (2) screws securing the plastic cover then remove the cover. Remove the two (2) screws securing the reagent syringe plunger clamp [2] then remove the clamp. Remove the thumbscrew [4] 5. securing the syringe and remove the syringe. Remove the seals from the syringe. Replacement Action VIDEO Install Seals and Reagent Syringe Steps Reference Note Video shows steps for the replacement of the Sample Syringe and Seals. However, steps are also applicable to the replacement of the Reagent Syringe and Seals. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Seals and Reagent Syringe 1. 2. 3. Install the seals on the syringe. Install the syringe [3] using the thumbscrew. Install the reagent syringe plunger clamp[2] using two (2) 4. 5. Prepare for Operation 1. 2. 3. 4. 5. screws. Connect the tubing [1] from the probe to the top of the reagent syringe and from the valve to the side of the reagent syringe. Install the plastic cover using two (2) screws. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G2.03 Syringe, Reagent Version - 204738-101_4923_1 List/Part Numbers List/Part Number Description 2-89398-02 Syringe, reagent, complete Inspect tools for damage, ensure calibration is not expired and replace if necessary. G2.03 Syringe, Reagent Time Required 18 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Open the left front door. Remove Reagent Syringe Note 1. 2. 3. 4. Use absorbent material to protect reagent syringe drive against spills. Unscrew the tubing [1] from the side and top of the reagent syringe. Remove the two (2) screws securing the plastic cover then remove the cover. Remove the two (2) screws securing the reagent syringe plunger clamp [2] then remove the clamp. Remove the thumbscrew [4] securing the syringe [3] then remove the syringe. Replacement Action Steps Install Reagent Syringe 1. 2. 3. 4. Install the syringe [3] using the thumbscrew. [4] Install the reagent syringe plunger clamp [2] using two (2) screws. Install the plastic cover using two (2) screws. Connect the tubing [1] from the probe to the top of the reagent syringe and from the valve to the side of the reagent syringe. Reference Prepare for Operation 1. 2. 3. 4. 5. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G3.01 Valve, Reagent Sample Syringe Dispense Version - 204738-101_4924_1 List/Part Numbers List/Part Number Description 7-202518-01 Valve, Reagent / Sample, CKD Inspect tools for damage, ensure calibration is not expired and replace if necessary. G3.01 Valve, Reagent/Sample Syringe Dispense Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. Remove Reagent/Sample Syringe Dispense Valve Note 1. 2. 3. 4. Place an absorbent towel under valve while removing to catch any water that maybe still in the tubing. Loosen the one (1) screw securing the Pressure Monitor sensor bracket [1] above the valve being removed and allow the bracket to float. Unscrew the fitting on the valve being removed to disconnect the tubing [2] from the NC port. Unscrew the compression nut [3] from the fitting screwed into the manifold. Loosen the one (1) screw [4] securing the valve mounting bracket to the metal brace. Remove Reagent/Sample Syringe Dispense Valve (continued) 1. 2. 3. 4. 5. 6. Lift the valve and bracket up and remove the bracket from the metal brace. Disconnect the appropriate plug: • SWV (Sample) • R1WV (R1) • R2WV (R2) Disconnect the tubing [5] from the NO port needle (rear of valve). Remove the one (1) screw [6] securing the valve to the mounting bracket and remove the valve from the mounting bracket. Remove the fitting [7] from the NO port. Unscrew the compression nut [8], with the tube and previously removed compression nut from the IN (COM) port. Replacement Action Steps Install Reagent/Sample Syringe Dispense Valve 1. 2. 3. 4. 5. Proper orientation of valve is with the COM port facing down. Install the mounting bracket to the valve using one (1) screw [6]. Install the pipe sealant tape on the threads of the compression fitting [8] that screws in the valve and screw the fitting into the IN (COM) port of the valve. Install the pipe sealant tape on the threads of the fitting [7] for the NO port and screw in the port of the valve. Connect the appropriate valve plug: • SWV Reference 6. Install Reagent/Sample Syringe Dispense Valve (continued) 1. 2. 3. 4. 5. • R1WV • R2WV Slide the tubing [5] onto the needle of the fitting for the NO port. Install the valve in position, ensuring that the compression nut [3] is aligned with the fitting screwed into the manifold. Tighten the compression nut [3] onto the fitting attached to the manifold. Secure the valve mounting bracket to the metal brace by tightening one (1) screw [4]. Screw in the fitting with tubing [2] from the syringe to the NC port of the valve. Tighten the one (1) screw [1] for the Pressure Monitor sensor bracket. Prepare for Operation 1. 2. 3. 4. 5. 6. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. G3.02 P.M. Sensor Assy Version - 204738-101_4925_2 List/Part Numbers List/Part Number Description 7-86320-01 PM Sensor Assy Inspect tools for damage, ensure calibration is not expired and replace if necessary. G3.02 P.M. Sensor Assy Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the supply center (left front) door. 2. VIDEO Remove P.M. Sensor Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove P.M. Sensor 1. 2. Use absorbent material to protect the sample syringe drive and the P.M. Sensor against spills. Disconnect the tubing from the top and bottom of the P.M. Sensor. 3. Loosen the one (1) screw [1] holding the sensor bracket to the frame, then lift sensor and bracket up and out to remove the sensor bracket. Disconnect sensor connector [2] from the P.M. Sensor. Remove the two (2) screws [3] securing the sensor and remove the sensor off of the bracket. 4. 5. Replacement Action Steps VIDEO Install P.M. Sensor Note Reference Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install P.M. Sensor 1. 2. 3. 4. Secure the P.M. Sensor on to the bracket by using two (2) screws. [3] Connect the sensor connector [2] to the P.M. Sensor. Install the sensor and the bracket to the frame using one (1) screw. [1] Connect the tubing to the top and bottom of the P.M. Sensor. Prepare for Operation 1. 2. 3. 4. 5. 6. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6053 Probe Water Wash. Check for leaks. Close the left front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6053 - Probe Water Wash Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. K1.01 Vacuum Pump Version - 204738-101_4926_1 List/Part Numbers List/Part Number Description 7-205194-01 VACUUM PUMP Inspect tools for damage, ensure calibration is not expired and replace if necessary. K1.01 Vacuum Pump Time Required 30 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door to access the vacuum pump assembly. Remove the right front access panel. 2. 3. Remove Vacuum Pump Assembly 1. 2. 3. 4. 5. 6. Reference Label the four tubes attached to the vacuum pumps. Clamp each tube. Disconnect each tube from the pump. Disconnect the vacuum pump power [1] cable located above the vacuum pump assembly. Remove four screws securing the pump assembly to the frame. Remove the pump assembly from the processing module. Replacement Action Steps Install Vacuum Pump 1. 2. 3. 4. Prepare for Operation 1. Reference Install the pump assembly to the frame using four screws. Connect vacuum pump power cable. Install the tubes. Release clamps from tubes. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 2. 3. 4. 5. 6. proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close right front door. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. K1.02 Diaphragm, Vac. Pump Version - 204738-101_4927_1 List/Part Numbers List/Part Number Description 2-94796-02 Diaphragm, Vacuum Pump New Style Inspect tools for damage, ensure calibration is not expired and replace if necessary. K1.02 Diaphragm, Vac. Pump Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door to access the vacuum pump assembly. Remove the right front access panel. 2. 3. Remove Vacuum Pump Diaphragm 1. 2. 3. 4. Label the four tubes attached to the vacuum pumps. Clamp each tube. Disconnect each tubes from the pump. Remove four screws securing diaphragm housing cover, and remove the cover. Reference Remove Vacuum Pump Diaphragm (continued) 1. Unscrew the diaphragm counterclockwise (CCW) to remove it from the vacuum pump. Replacement Action Steps Install Vacuum Pump Diaphragm 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. Reference Screw the diaphragm clockwise (CW) on to the rod. Note The diaphragm should only be hand tightened. Install the diaphragm housing cover using four (4) screws. Install the tubes. Release clamps from tubes. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from Stopped to System Operations Manual, Section 5: Operating instructions, Ready, perform the Startup procedure. Subsection: System startup, pause, and shutdown. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close right front door. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. K1.03 Sheet Valve, Vac. Pump Version - 204738-101_4928_1 List/Part Numbers List/Part Number Description 2-94797-02 Sheet Valve for Vacuum Pump New Style Inspect tools for damage, ensure calibration is not expired and replace if necessary. K1.03 Sheet Valve, Vac. Pump Time Required 10 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door to access the vacuum pump assembly. Remove the right front access panel. 2. 3. Remove Vacuum Pump Sheet Valve 1. 2. 3. 4. 5. 6. Label the tubes attached to the vacuum pumps. Clamp each tube. Disconnect each tube from the pump. Remove four (4) screws securing diaphragm housing cover, and remove the cover. Separate the two cover pieces. Remove the sheet valve; note the position. Reference Replacement Action Steps Install Vacuum Pump Sheet Valve 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. 5. Reference Install the sheet valve as noted. Assemble the two cover pieces. Install the diaphragm housing cover using four (4) screws. Install the tubes. Release clamps from tubes. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from Stopped to System Operations Manual, Section 5: Operating instructions, Ready, perform the Startup procedure. Subsection: System startup, pause, and shutdown. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close right front door. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks Expected Results N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. K1.04 O-ring, Vacuum Pump Version - 204738-101_4929_1 List/Part Numbers List/Part Number Description 2-94798-02 O-Ring for Vacuum Pump New Style Inspect tools for damage, ensure calibration is not expired and replace if necessary. K1.04 O-ring, Vacuum Pump Time Required 10 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door to access the vacuum pump assembly. Remove the right front access panel. 2. 3. Remove Vacuum Pump O-Ring 1. 2. 3. 4. 5. 6. Label the four tubes attached to the vacuum pumps. Clamp each tube. Disconnect each tube from the pump. Remove four (4) screws securing diaphragm housing cover, and remove the cover. Separate the two (2) cover pieces. Remove the O-ring. Replacement Action Steps Install Vacuum Pump 1. O-Ring 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. 5. Reference Install the O-ring. Assemble the two (2) cover pieces. Install the diaphragm housing cover using four (4) screws. Install the tubes. Release clamps from tubes. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Close right front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. K2.01 Degasser Version - 204738-101_4930_1 List/Part Numbers List/Part Number Description 2-89161-05 New Style Degasser Unit (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. K2.01 Degasser Time Required 30 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status.. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the left rear cover. 2. Remove Degasser 1. 2. 3. 4. Reference Clamp the tubing [1] to the degasser, and disconnect it. Disconnect the DEG and DGS cables from the degasser. Remove the two (2) screws securing the degasser assembly to the processing module frame. Remove the degasser by sliding it out from the rear of the processing module. Replacement Action Steps Install Degasser 1. Slide the degasser in the processing module Reference 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. from the rear and secure with two (2) screws. Connect the DEG and DGS cables to the degasser. Connect the tubing to the degasser and remove the clamps. Secure the frame brace with four (4) screws. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2132 Flush Water Lines. Check for leaks. Install the right rear cover and secure with two screws. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. K2.02 Degasser Canister Version - 204738-101_4931_1 List/Part Numbers List/Part Number Description 2-89161-06 Degasser, with improvements 7-93668-02 New Style Degasser Canister, With Tubing Inspect tools for damage, ensure calibration is not expired and replace if necessary. K2.02 Degasser Canister Time Required 60 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the left rear cover. 2. Remove Degasser 1. 2. 3. 4. Remove Canister 1. 2. Clamp the tubing [1] to the degasser, and disconnect it. Disconnect the DEG and DGS cables from the degasser. Remove the two (2) screws securing the degasser assembly to the processing module frame. Remove the degasser by sliding it out from the rear of the processing module. Remove the two (2) screws [1] securing the canister to the rear of the degasser unit. Disconnect the vacuum tubing [2] from the front of the canister. Reference 3. 4. 5. Remove the two (2) screws [3] securing the canister to the bracket. Remove the two (2) screws [4] securing the canister bracket to the degasser unit base plate. Slide the canister out of the degasser unit. Replacement Action Steps Install Canister 1. 2. 3. 4. 5. Slide the canister into the degasser unit. Install the two (2) screws [1] that secure the canister to the rear of the degasser unit. Connect the vacuum tubing [2] to the front of the canister. Install the two (2) screws [3] that secure the canister to the canister bracket. Install the two (2) screws [4] that secure the canister bracket to the degasser unit base plate. Reference Install Degasser 1. 2. 3. Prepare for Operation 1. 2. 3. 4. 5. 6. Slide the degasser in the processing module from the rear and secure with two (2) screws. Connect the DEG and DGS cables to the degasser. Connect the tubing [2] to the degasser and remove the clamps. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2132 Flush Water Lines. Check for leaks. Install the left rear cover. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. L1.01 Main Power Supply Version - 204738-101_5007_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. L1.01 Main Power Supply Time Required 90 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Lifting Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker). Remove the left rear cover. 2. Remove Main Power Supply 1. 2. 3. 4. 5. Disconnect power cord. Remove three (3) screws on front of the power supply to the processing module frame. Disconnect the cable from the power supply to the transformer. Note Note the selected voltage. Disconnect the following cables from the side of the power supply: Note Start from the back of the analyzer and continue to slide out the power supply while removing the cables. • J4/P4 • J1/P1 • J6/P6 • CPUM • CNBD • UC • +12 V • CPU • Lamp cables • AUTO • ON/OFF • Thermostat • TBC • AC100V • +24V1 • +24V2 • STRD • AC • Transformer • RSH Jumper cable-make note of the cable connection configuration. Remove the power supply from the analyzer and prepare it to Reference be packed in the box containing the new one. Replacement Action Steps Install Main Power Supply 1. 2. 3. 4. 5. 6. 7. Connect the RSH cable according to instrument configuration and as noted during the removal process Slide the power supply into the processing module. Connect the cables in the following order: • STRD • AC • Transformer • AC100V • +24V1 • +24V2 • ON/OFF • AUTO • Thermostat • TBC • CPUM • CNBD • UC • +12 V • CPU • Lamp cable • J4/P4 • J1/P1 • J6/P6 Replace three (3) screws on the front of the power supply to the processing module frame. Connect the cable from the power supply to the transformer. Install left front rear cover. Connect the power cord. Reference Prepare for Operation 1. 2. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT System Note The system control center power must be ON Operations Manual, Section 5: Operating instructions, Subsection: System startup, prior to turning on the processing module to pause, and shutdown. ensure proper initialization. Perform the Startup procedure to change the status of the processing module and sample handler from Stopped to Ready. Verification Procedures Order VP Description 1 5712 - Voltages Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.01 Card Cage without Boards Version - 204738-101_4932_1 List/Part Numbers List/Part Number Description 7-206858-01 Card cage without Bds, c4 7-206859-01 Card Cage, With Bds, c4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.01 Card Cage without Boards Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Disconnect the power cord. Remove the right rear cover. Open the right front door. Remove front access panel. 2. 3. 4. 5. Disconnect CPU Cables 1. 2. Remove Main Card Cage 1. 2. From the rear of the processing module, disconnect the two cables labeled CPU and CPUM from the main power supply. Remove cables CPU1, CPU2, CPU3, CPU4, CPU5, CPU6, CPU7, CPU8, CPU9, CPU10, CPU11, CPU12, and CPU14 from the rear of the card cage. Remove the two (2) screws located at the front of the card cage. Remove the card cage. Reference Remove Boards and 1. Cables 2. 3. 4. 5. 6. Loosen the two (2) captive thumbscrews securing the card cage door and open the door. Disconnect the MECH LED cables from the SM AC/DC Bd, [4] SMC Bd, [3] and the AC/DC Controller Bd [5] Unplug the ethernet cable from the front of the CPU board. [6] Remove the screws on the boards located by the ejectors. Open the ejectors to remove all of the boards from the card cage. Remove the MECH LED board from the plastic standoffs on the card cage front cover. Replacement Action Steps Install Boards and Cables 1. 2. 3. 4. 5. Install all the boards into the card cage and secure using the screws by the ejectors. Mount the MECH LED board on the plastic standoffs of the card cage front cover. Connect the MECH LED cables to the SMC board, [3] SM AC/DC board, [4] and the AC/DC controller board. [5] Connect the ethernet cables to the CPU board. [6] Close the card cage front cover and secure using two (2) captive thumbscrews. Install Main card Cage 1. From the front of the processing module, slide the card cage in the processing module and secure it with the screws. Connect CPU Cables 1. Connect the two power cables labeled CPU and CPUM to the main power supply. On the rear of the card cage backplane connect the cables: • CPU1 • CPU2 • CPU3 • CPU4 • CPU5 • CPU7 • CPU8 • CPU9 • CPU10 • CPU11 • CPU12 • CPU14 2. Prepare for Operation 1. 2. Reference Install the left rear cover and secure using four (4) captive screws. Install the front access panel. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 3. 4. 5. 6. Close all doors. Connect the processing module to the AC supply power. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.02 Bd., DAQ Version - 204738-101_4933_1 List/Part Numbers List/Part Number Description 7-93181-04 DAQ board, RoHS compliant 7-93181-05 DAQ board NS, RoHS compliant Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.02 Bd., DAQ Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 2. Preparation 1. 2. 3. Remove DAQ Board 1. 2. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) thumbscrews securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the DAQ board [1] located inside the card cage. Locate the DAQ board and loosen the two locking screws above the upper ejector and below the lower ejector [1]. Press up on the upper ejector and down on the lower ejector and remove the DAQ board from the card cage. Replacement Action Steps Install DAQ Board 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. 5. Reference With the upper ejector in the up position and the lower ejector in the down position, slide the DAQ board into the correct slot position until the ejectors engage in the upper and lower supports. Press down gently on the upper and up on the lower ejectors to insert the DAQ board into the back plane. Secure the DAQ board with the two (2) locking screws above the upper ejector and below the lower ejector. Secure the card cage front cover with the two thumbscrews. Secure the right front access panel with the four screws. Close all doors. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Diagnostics, Modules, 6001 Save/Restore DAQ Data and load the step table data from CPU DAQ. Select System, Maintenance, As Needed, 6070 Daily Maintenance. Note 6070 Daily Maintenance initializes the ICT channels in SRAM on the DAQ board. If Daily Maintenance is not performed after installing a new DAQ board or during new system installation, one or more of the three ICT channels could be uninitialized. This will result in a zero A/D read and error 1600 Unable to calculate result, mV reading outside measurable range of ICT Unit. Verification Procedures Order VP Description VP Detail / Note Expected Results 2 6070 - Daily Maintenance 1 6001 - Save/Restore DAQ Data N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.03 Bd., SMC Version - 204738-101_4934_1 List/Part Numbers List/Part Number Description 7-202599-01 Bd., SMC Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.03 Bd., SMC Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 2. Preparation 1. 2. 3. Remove SMC Board 1. 2. 3. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two thumbscrews securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the SMC board [1] located inside the card cage. Locate the SMC board and loosen the two locking screws above the upper ejector and below the lower ejector [1]. Disconnect the MECH LED board cable SMC1 from the SMC board. Press up on the upper ejector and down on the lower ejector and remove the SMC board from the card cage. Replacement Action Steps Install SMC Board 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. Reference With the upper ejector in the up position and the lower ejector in the down position, slide the SMC board into the correct slot position until the ejectors engage in the upper and lower supports. Press down gently on the upper and up on the lower ejectors to insert the SMC board into the back plane. Secure the SMC board with the two locking screws above the upper ejector and below the lower ejector. Connect the MECH LED cable SMC1 to the SMC board. Secure the card cage front cover with the two (2) thumbscrews. Secure the right front access panel with the four (4) screws. Close all doors. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Maintenance, As Needed, 6070 Daily Maintenance. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.04 Bd., AC/DC Controller Version - 204738-101_4935_2 List/Part Numbers List/Part Number Description 7-93183-04 Bd, AC DC Controller (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.04 Bd., AC/DC Controller Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. Preparation 1. 2. 3. Remove AC/DC Controller Board 1. 2. 3. The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) thumbscrews securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the AC/DC Controller board [1] located inside the card cage. Locate the AC/DC Controller board and loosen the two locking screws above the upper ejector and below the lower ejector [1]. Disconnect the MECH LED board cable AC/DC1 from the AC/DC Controller board. Press up on the upper ejector and down on the lower ejector and remove the AC/DC Controller board from the card cage. Replacement Action Steps Install AC/DC Controller Board 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Reference With the upper ejector in the up position and the lower ejector in the down position, slide the AC/DC Controller board into the correct slot position until the ejectors engage in the upper and lower supports. Ensure switches S2-1 & S2-8 are ON for a c4000. Press down gently on the upper and up on the lower ejectors to insert the AC/DC Controller board into the back plane. Secure the AC/DC Controller board with the two locking screws above the upper ejector and below the lower ejector. Connect the MECH LED cable AC/DC1 to the AC/DC Controller board. Secure the card cage front cover with the two (2) thumbscrews. Secure the right front access panel with the four (4) screws. Close all doors. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform P-601 Reagent Supply Center Temperature Detector Adjustment. Perform P-604 Internal Instrument Temperature Detector Adjustment. Perform P-605 Water Bath Temperature Detector Adjustment. Select System, Maintenance, As Needed, 6070 Daily Maintenance. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. ARCHITECT c4000 Service and Support Manual (Version 204733-112) • © 2009, 2011 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. null VP Detail / Note Expected Results M1.05 Bd., CPU Version - 204738-101_4936_1 List/Part Numbers List/Part Number Description 7-205719-01 CPU BOARD C4 7-205719-02 CPU c4000, New Style Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.05 Bd., CPU Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 2. Preparation 1. 2. 3. Remove CPU Board 1. 2. 3. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) thumbscrews securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the CPU board [1] located inside the card cage. Locate the CPU board and loosen the two locking screws above the upper ejector and below the lower ejector [1]. Disconnect the ethernet cable [2] from the front of the CPU board. Press up on the upper ejector and down on the lower ejector and remove the CPU board from the card cage. Replacement Action Steps Install CPU Board 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. 5. Reference With the upper ejector in the up position and the lower ejector in the down position, slide the CPU board into the correct slot position until the ejectors engage in the upper and lower supports. Press down gently on the upper and up on the lower ejectors to insert the AC/DC Controller board into the back plane. Secure the CPU board with the two (2) locking screws above the upper ejector and below the lower ejector. Connect the ethernet cable to the front of the CPU board. Secure the card cage front cover with the two (2) thumbscrews. Secure the right front access panel with the four (4) screws. Close all doors. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Diagnostics, Modules, 6001 Save/Restore DAQ Data and load the step table data from CPU DAQ. Select System, Maintenance, As Needed, 6070 Daily Maintenance. Verification Procedures Order VP Description 2 6070 - Daily Maintenance 1 6001 - Save/Restore DAQ Data N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.06 Bd., Mech LED Version - 204738-101_4937_2 List/Part Numbers List/Part Number Description 7-202564-01 Bd., Mech LED,m 16 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.06 Bd., Mech LED Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 2. Preparation 1. 2. 3. Remove MECH LED 1. Board 2. 3. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) captive thumbscrews securing the card cage front cover and open the cover. Locate the MECH LED board on the back of the card cage cover. [1] Disconnect the three (3) cables LED1, LED2 and LED3 from the top of the MECH LED Board. Remove the MECH LED board from the plastic standoffs on the card cage cover. Replacement Action Steps Reference Install MECH LED Board 1. 2. Prepare for Operation 1. 2. 3. 4. 5. 6. Install the MECH LED board on the plastic standoffs, on the card cage. Connect the three (3) cables LED1, LED2 and LED3 to the top of the MECH LED Board. Note: If installing a New Style MECH LED Board, refer to the Service Manual, Section 2 Troubleshooting/Card Cage/New Style MECH LED for a diagram of the board and LED reference. Close all doors. Power ON the processing module. Note: The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Verify LED operation on the MECH LED Board. (LEDs will light up depending on sensor status.) Close the card cage front cover and secure using two (2) captive thumbscrews. Secure the right front access panel with the four (4) screws. Close the right front door. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 G1 - Verify the Repair according to released Operation and Service Procedures. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.07 Fan Assy., Main Card Cage Version - 204738-101_4938_2 List/Part Numbers List/Part Number Description 7-93188-01 Fan assy, main card cage Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.07 Fan Assy., Main Card Cage Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Disconnect the power cord. Remove the right rear cover. Open the right front door. Remove front access panel. 2. 3. 4. 5. Remove Card Cage Fan Assembly Note 1. 2. 3. Fan airflow direction is UP into the card cage. Loosen the two (2) captive thumbscrews securing the card cage door and open the door. Disconnect the power cable [3] to fan assembly. Remove the two screws on front of fan assembly [4] and pull the assembly straight out of the card cage. Replacement Action Steps Install Card Cage Fan Assembly Note 1. Verify fan airflow direction is UP into the card cage. 6. Slide the fan assembly [4] straight in the card cage and secure using two (2) screws. Note Ensure ground wire is installed with screw on right side of the fan assembly. Connect the power cable [3] to fan assembly. Close the card cage front cover and secure using two (2) captive thumbscrews. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verify that the fan blades are turning. 1. 2. Install the front access panel. Close all doors. 2. 3. 4. 5. Prepare for Operation Reference For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. ARCHITECT c4000 Service and Support Manual (Version 204733-104) • © 2009, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200, c16000, ci16200, ci4100 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. Verification Procedures Order VP Description VP Detail / Note 1 G14 - Fan Verify that the fan is rotating and that the air flows in the correct direction. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results M1.08 Bd., Pressure Monitor Version - 204738-101_4939_4 List/Part Numbers List/Part Number Description 7-900040-01 BOARD, PM, C4000 7-900040-02 Board, PM, c16000 and c4000 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.08 Bd., Pressure Monitor Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 2 Preparation 1. 2. 3. Check Switch Setting Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) thumbscrews, securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the P.M. Board [1] located inside the card cage. Note This information is for the -107 board only. The instrument must be at software version 7.0 or higher. 1.Switches 1, 2, 3, 4, 6, & 8 will be in the OFF position from the factory. Remove PM Board 1. 2. Locate the PM board and loosen the two (2) locking screws above the upper ejector and below the lower ejector [1]. Press up on the upper ejector and down on the lower ejector and remove the PM Board from the card cage. Replacement Action Steps Install PM Board 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. Reference With the upper ejector in the up position and the lower ejector in the down position, slide the PM board into the correct slot position until the ejectors engage in the upper and lower supports Press down gently on the upper and up on the lower ejectors to insert the PM board into the back plane. Secure the PM board with the two (2) locking screws above the upper ejector and below the lower ejector. Secure the card cage front cover with the two (2) thumbscrews. Secure the right front access panel with the four (4) screws. Close all doors. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Diagnostics, Modules, 3805 Pressure Monitor Test. Verification Procedures Order VP Description 1 G1 - Verify the Repair according to released Operation and Service Procedures. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.09 Flash Memory, CPU Version - 204738-101_4940_1 List/Part Numbers List/Part Number Description 7-206800-01 Flash Memory, CPU c4 7-206800-02 Flash Memory, New Style CPU c4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.09 Flash Memory, CPU Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 1. Preparation 1. 2. 3. Remove CPU Board 1. 2. 3. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) thumbscrews securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the CPU Board [1] located inside the card cage. Locate the CPU board and loosen the two locking screws above the upper ejector and below the lower ejector [1]. Disconnect the Ethernet cable [2] from the front of the CPU board. Press up on the upper ejector and down on the lower ejector and remove the CPU board from the card cage. Remove Flash Memory 1. Slide the ejector tab [1] on the side of the flash memory socket. Note This ejects the flash memory for removal. Replacement Action Steps Install Flash Memory 1. Slide the flash memory into the socket until it latches in place. Install CPU Board With the upper ejector in the up position and the lower ejector in the down position, slide the CPU board into the correct slot position until the ejectors engage in the upper and lower supports. Press down gently on the upper and up on the lower ejectors to insert the CPU board into the back plane. Secure the CPU board with the two (2) locking screws above the upper ejector and below the lower ejector. Connect the Ethernet cable to the front of the CPU board. Secure the card cage front cover with the two (2) thumbscrews. Secure the right front access panel with the four (4) screws. 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. Close all doors. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing Reference For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 4. 5. module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Diagnostics, Modules, 6001 Save/Restore DAQ Data and load the step table data from CPU DAQ. Select System, Maintenance, As Needed, 6070 Daily Maintenance. Verification Procedures Order VP Description 2 6070 - Daily Maintenance 1 6001 - Save/Restore DAQ Data N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.10 Card Cage with Boards Version - 204738-101_4941_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.10 Card Cage with Boards Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Disconnect the power cord. Remove the right rear cover. Open the right front door. Remove front access panel. 2. 3. 4. 5. Disconnect CPU Cables 1. 2. Remove Main Card Cage 1. Reference From the rear of the processing module, disconnect the two cables labeled CPU and CPUM from the main power supply. Remove cables CPU1, CPU2, CPU3, CPU4, CPU5, CPU6, CPU7, CPU8, CPU9, CPU10, CPU11, CPU12, and CPU14 from the rear of the card cage. Remove the two (2) screws located at the front of the card cage and pull the card cage out of the system. Replacement Action Steps Install Main card 1. From the front of the processing module, Reference Cage Connect CPU Cables slide the card cage in the processing module and secure using two (2) screws at the bottom, front of the card cage. 1. 2. Prepare for Operation 1. 2. 3. 4. 5. 6. Connect the two power cables labeled CPU and CPUM to the main power supply. On the rear of the card cage backplane connect the cables: a. CPU1 b. CPU2 c. CPU3 d. CPU4 e. CPU5 f. CPU7 g. CPU8 h. CPU9 i. CPU10 j. CPU11 k. CPU12 l. CPU14 Install the left rear cover and secure using For information on the startup procedure, refer to the ARCHITECT four (4) captive screws. System Operations Manual, Section 5: Operating instructions, Install the front access panel. Subsection: System startup, pause, and shutdown. Close all doors. Connect the processing module to the AC supply power. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M1.11 Bd., SM AC/DC Controller Version - 204738-101_4942_1 List/Part Numbers List/Part Number Description 7-202601-01 Bd., SM AC/DC Controller 7-202601-02 Bd., SM ACDC Controller Inspect tools for damage, ensure calibration is not expired and replace if necessary. M1.11 Bd., SM AC/DC Controller Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. 2. Preparation 1. 2. 3. Remove SM AC/DC Board 1. 2. 3. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) thumbscrews securing the card cage front cover and open the cover. Note There is a diagram on the card cage cover that shows the position of the SM AC/DC Board [1] located inside the card cage. Locate the SM AC/DC board and loosen the two locking screws above the upper ejector and below the lower ejector [1]. Disconnect the MECH LED board cable SM AC/DC1 from the SM AC/DC board. Press up on the upper ejector and down on the lower ejector and remove the SM AC/DC board from the card cage. Replacement Action Steps Install SM AC/DC Board 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. Reference With the upper ejector in the up position and the lower ejector in the down position, slide the SM AC/DC board into the correct slot position until the ejectors engage in the upper and lower supports. Press down gently on the upper and up on the lower ejectors to insert the SM AC/DC board into the back plane. Secure the SM AC/DC board with the two (2) locking screws above the upper ejector and below the lower ejector. Connect the MECH LED cable SM AC/DC1 to the SM AC/DC board. Secure the card cage front cover with the two (2) thumbscrews. Secure the right front access panel with the four (4) screws. Close all doors. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select System, Maintenance, As Needed, 6070 Daily Maintenance. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M2.01 Bd., Mixer Driver Version - 204738-101_4943_1 List/Part Numbers List/Part Number Description 2-89035-05 BD, Mixer Driver (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. M2.01 Bd., Mixer Driver Time Required 45 min Tools/Materials Standard Tool Kit Multimeter Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Unplug the processing module from the AC supply power. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) screws [1] on the Driver Box door and open the door. Note Lift the door up slightly to clear the screws located in front of the driver box assembly. 2. 3. 4. 5. Reference Remove Mixer Driver Board 1. 2. Disconnect plugs labeled • STRD1 • STRD2, STRD3 • STRD4 Remove the board from the four (4) standoffs. Replacement Action Steps Install Mixer Driver Board 1. 2. Place the board onto the four (4) standoffs. Ensure the board is locked into place. Connect plugs • STRD1 Reference • STRD2 • STRD3 • STRD4 to connectors • CNN1/PJ1 • CNN2/PJ2 • CNN3/PJ3 • CNN4/PJ4 Prepare for Operation 1. 2. 3. 4. 5. 6. Close the Driver Box door and secure with For information on the startup procedure, refer to the ARCHITECT two (2) screws. System Operations Manual, Section 5: Operating instructions, Install the right front access panel using four Subsection: System startup, pause, and shutdown. (4) screws. Close the right front door. Connect the processing module AC supply power. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 3126 - Mixer Vibration Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M2.02 Bd., SMD/5 Phase Motor Driver Version - 204738-101_4944_1 List/Part Numbers List/Part Number Description 7-205178-01 BD, SMD/5 PHASE MOTOR DRIVER, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M2.02 Bd., SMD/5 Phase Motor Driver Time Required 45 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Unplug the processing module from the AC supply power. Open the right front door. Remove the right front access panel by removing the four (2) screws. Loosen the two (2) screws [1] on the Driver Box door and open the door. Note Lift the door up slightly to clear the screws located in front of the driver box assembly. 2. 3. 4. 5. Reference Remove SMD/5 Phase Board 1. 2. Remove plugs 1-22 from the connectors. Note Squeeze tabs inward on plug SMD1 to release the plug from the connector on the board. DO NOT push the tabs outward or it will break. [1] Remove the board from the nine (9) standoffs. [1] Replacement Action Steps Install SMD/5 Phase Board 1. 2. Install the board on the nine (9) standoffs. Connect plugs 1-22 to the connectors. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. Close the Driver Box door and secure with For information on the startup procedure, refer to the ARCHITECT two (2) screws. System Operations Manual, Section 5: Operating instructions, Install the right front access panel using four Subsection: System startup, pause, and shutdown. (4) screws. Close the right front door. Connect the processing module AC supply power. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M2.03 Bd., AC/DC Driver Version - 204738-101_4945_1 List/Part Numbers List/Part Number Description 7-205350-01 BD, ACDC DRIVER, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M2.03 Bd., AC/DC Driver Time Required 45 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Unplug the processing module from the AC supply power. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) screws [1] on the Driver Box door and open the door. Note Lift the door up slightly to clear the screws located in front of the driver box assembly. 2. 3. 4. 5. Reference Remove AC/DC Driver Board 1. 2. Disconnect plugs 1-13. Note Squeeze tabs inward on plug SMD1 to release the plug from the connector on the board. DO NOT push the tabs outward or it will break. Remove the AC/DC Driver board from the nine (9) standoffs. Replacement Action Steps Install AC/DC Driver 1. Board 2. Install the AC/DC Driver Board on the nine (9) standoffs. Connect plugs labeled 1-13. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. Close the Driver Box door and secure with For information on the startup procedure, refer to the ARCHITECT two (2) screws. System Operations Manual, Section 5: Operating instructions, Install the right front access panel using four Subsection: System startup, pause, and shutdown. (4) screws. Close the right front door. Connect the processing module AC supply power. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M2.04 Bd., Motor Driver Version - 204738-101_4946_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. M2.04 Bd., Motor Driver Time Required Not Assessed Tools/Materials Not Assessed WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Note Function for 5 - phase vertical drive for the sample arm. Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Loosen the four (4) captive screws securing the left rear cover and remove the cover. Unplug the processing module from the AC supply power. Remove two (2) screws securing the bracket to the processing module frame. Remove one (1) screw securing bracket to the right side of the processing module. Slide the driver box assembly out of the processing module. Remove one (1) screw securing the cover to the driver box assembly and open up the assembly. 2. 3. 4. 5. 6. 7. Reference Remove Motor Driver Board 1. 2. Disconnect cables DRV1, DRV2, DRV3 [1] from the motor driver board Remove two (2) screws [2] securing the board to the driver box frame. Replacement Action Steps Install Motor Driver Board 1. 2. Install the board using two (2) screws. [2] Connect cables DRV1, DRV2, and DRV3 [1] to the motor driver board. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. Connect the processing module AC supply For information on the startup procedure, refer to the ARCHITECT power. System Operations Manual, Section 5: Operating instructions, Power ON the processing module. Subsection: System startup, pause, and shutdown. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Install the cover to the driver box driver assembly using one (1) screw. Note When closing the driver box assembly, be careful not to pinch or crimp cables. Slide the driver box assembly into the processing module. Install the driver box assembly to the processing module using three (3) screws. Install the left rear cover using four (4) captive screws. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M3.01 Bd., ICT Preamp Version - 204738-101_4947_1 List/Part Numbers List/Part Number Description 2-89032-04 BD, ICT Preamp (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. M3.01 Bd., ICT Preamp Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the right front cover [C]. Disconnect the cable from the LUI board. [1] Remove the two (2) screws that secure the LUI board.[2] Remove screw [3] securing the Pre-amp module to the frame. 2. 3. 4. 5. Reference Remove ICT PreAmp Board 1. 2. 3. Loosen the two (2) screws securing the pre-amp enclosure and remove the cover. Disconnect the signal cable from the bottom of the pre-amp board. Remove screw securing the black ground to the instrument frame. Remove ICT PreAmp Board (continued) 1. 2. 3. Loosen the retaining nut on the pre-amp cable connector at the top of the enclosure. Remove the cable connector from the case. Remove the pre-amp board from the standoffs. Replacement Action Steps Install ICT Pre-Amp Board 1. 2. 3. 4. 5. 6. Install the pre-amp board onto the standoffs in the pre-amp enclosure. Install the pre-amp cable connector on the top side of the enclosure and secure with the retaining nut. Pass the ground cable through the opening at the bottom of the enclosure. Connect the signal cable at the bottom of the board. Attach the ground cable [3] to the enclosure. Note Make sure the ground cables do not obstruct the sliding path. Secure the pre-amp enclosure cover with two screws. Reference 7. Slide the ICT pre-amp module back into instrument and secure with the one screw. 8. Connect the ICT pre-amp cable. 9. Install the LUI board and connect the cable. 10. Install the right front cover. Prepare for Operation 1. 2. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G6 - Calibration Verify calibration per operator's manual. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M3.02 Bd., IREF Level Detector Version - 204738-101_4948_1 List/Part Numbers List/Part Number Description 2-93509-02 BD., IREF LVL Detect 7-93438-02 Slide mechanism, pipettor head Inspect tools for damage, ensure calibration is not expired and replace if necessary. M3.02 Bd., IREF Level Detector Time Required 40 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the front deck right front cover [C]. 2. Disconnect ICT Level Sensor Board 1. 2. 3. 4. 5. Disconnect the cable from the LUI board. [1] Remove the screws that secure the LUI board. [2] Remove the screw [3] that holds the ICT pre-amp module. Lift the ICT pre-amp module to gain access to the ICT Level sensor board. Loosen the screw from the sensor board cover; slide the cover slightly to the left and remove the cover. [4] Disconnect ICT Level Sensor Board (continued) 1. Disconnect cables ICT-L-1 and ILS-2 from the IREF level sensor board connectors CN1 and CN2. [1] Remove IREF Level Detector Board 1. Remove the IREF Level Detector Board from the four (4) standoffs. Replacement Action Steps Install IREF Level Detector Board 1. Install the IREF Level Detector Board onto the standoffs in the pre-amp enclosure. Connect IREF Level Detector Board. 1. Connect cables ICT-L-1 and ILS-2 to connectors CN1 and CN2 on the IREF Level Detector Board. Prepare for Operation 1. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6063 Flush ICT Module three times. While this is performed confirm the LEDs on the detector board function as indicated during ICT reference solution aspiration: a. Immediately before the ICT assembly drives down and aspirates ICT reference solution (out of the ICT reference solution cup) the LED1 flashes ON (read status) and LED2 flashes ON (ICT reference cup is filled). b. Immediately after the ICT aspirates ICT reference solution from the ICT reference solution cup and the assembly drives up 2. 3. 4. Reference 5. 6. 7. 8. out of the cup, LED1 flashes ON and LED2 stays OFF. [1] Note You must observe the ICT module and the board while performing the ICT flush procedure. [2] Install the detector board cover and secure with the screw. Slide the ICT assembly back in position. Note Make sure the ground cables do not obstruct the sliding path. Secure the ICT pre-amp module with the screw. [3] Install the right front cover. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 2 G6 - Calibration Verify calibration per operator's manual. 1 6063 - Flush ICT Module Check for leaks. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in Expected Results specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M4.01 Bd., LLS Version - 204738-101_4949_1 List/Part Numbers List/Part Number Description 7-205749-01 LLS BOARD, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M4.01 Bd., LLS Time Required 35 min Tools/Materials Standard Tool Kit Mutimeter WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Unplug the processing module from the AC power supply. Open the front doors. 2. 3. Remove Level Sense Board Assembly 1. 2. 3. Remove the top two (2) screws [1] and loosen the bottom two (2) screws [2] securing the right front access panel and remove panel. Remove the two (2) screws securing the Level Sense Board Cover. [3] Disconnect the following cables: • CN1 TLLS1 • CN2 TLLS2 • CN3 TLLS3 • TLLS-S/R • CN5 TLLS5 Reference Remove Level Sense Board Assembly (continued) 1. 2. Unscrew the cable [1] that secures the ground cables. Remove the board from the five (5) standoffs [2]. Replacement Action Steps Install LLS Board 1. 2. 3. Insert the ground cables onto the screw and tighten it. [1] Place the board on the five (5) standoffs. Connect the cables: Reference 4. 5. 6. Prepare for Operation 1. 2. 3. 4. • CN1 TLLS1 • CN2 TLLS2 • CN3 TLLS3 • TLLS-S/R • CN5 TLLS5 Install the cover of the Level Sense Board and secure it with the two (2) screws. Install the right front access panel. Close the front doors. Plug the processing module into the AC For information on the startup procedure, refer to the ARCHITECT power supply. System Operations Manual, Section 5: Operating instructions, Power ON the processing module. Subsection: System startup, pause, and shutdown. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform P-617 LLS Voltage Adjustment to verify voltage. Verification Procedures Order VP Description 1 3625 - Pipettors LLS Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. M5.01 Bd., CNN Version - 204738-101_4950_1 List/Part Numbers List/Part Number Description 7-205179-01 BD, CNN, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. M5.01 Bd.,CNN Time Required 45 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the front deck, front right cover [C]. Remove the rear left cover. 2. 3. Remove CNN Board 1. 2. 3. Loosen the two (2) screws that hold the CNN Board to the frame of the instrument and remove the board from them. [1] From the top of the instrument, lift the CNN board frame and gently slide it towards the back of the instrument. Note The CNN Board frame has two hooks that rest on the frame of the instrument towards the front, this section has to be lifted and pulled out as well to release the board. Disconnect all the cables from the back to Reference 4. the front. Note To remove some of the connectors you have to squeeze the edges while for others you need to open them to release the connector. Remove the CNN board from the instrument. Replacement Action Steps Install CNN Board 1. 2. 3. Slide the CNN Board to the instrument. Secure the board; make sure that the hooks toward the front and the orifices for the screws are sitting properly. Connect the cables from the back to the front as follows: a. CNN2 b. CNN3 c. CNN24 d. CNN7 e. CNN37 f. CNN38 g. CNN18 h. CNN23 i. CNN14 j. CNN13 k. CNN27 l. CNN17 m. CNN28 n. CNN15 o. CNN22 p. CNN25 q. CNN26 r. CNN29 s. CNN30 t. CNN11 u. CNN19 v. CNN16 w. CNN12 x. CNN21 y. CNN20 z. CNN35 aa. CNN33 ab. CNN36 Reference 4. 5. 6. Prepare for Operation 1. 2. ac. CNN5 ad. CNN6 ae. CNN1 af. CNN31 ag. CNN4 ah. CNN32 ai. CNN34 Note Make sure all connectors are properly secured. Secure the CNN board with the screws. Install the rear left cover. Install the Front right cover [C]. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.01 Sensor, Trigger, Reaction Carousel Version - 204738-101_4951_1 List/Part Numbers List/Part Number Description 7-205307-01 OPT TRG SNSR, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.01 Sensor, Trigger, Reaction Carousel Time Required 25 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the two (2) screws securing the rear right cover [K] and remove the cover. 2. Remove Trigger Sensor 1. 2. 3. Disconnect cable from the OPTTRG connector. Remove the OPTTRG plug from the connector plate. Remove the two (2) screws [1] securing the trigger sensor [2] to the mounting bracket and remove the sensor. Reference Replacement Action Steps Install Trigger Sensor 1. 2. 3. 4. 5. 6. Install the trigger sensor [2] on the mounting bracket using two (2) screws. Insert the OPTTRG connector into the connector plate. Connect the cable to the OPTTRG connector. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Select the Optics/Temperature category from the Diagnostics screen under the System icon. Perform 1001 Optics Trigger Sensor Check. Note Adjustment of the sensor position is made at the end of the mounting bracket opposite the sensor. Reference Prepare for Operation 1. 2. Install the rear right cover [K] using two (2) screws. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.02 Cuvette Wiper Version - 204738-101_4952_2 List/Part Numbers List/Part Number Description 09D42-02 Cuvette Wiper Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.02 Cuvette Wiper Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the right rear cover [K] and the right, center rear cover [J]. 2. VIDEO Remove Cuvette Wipers Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Cuvette 1. Wipers 2. 3. 4. 5. Remove a cuvette segment from the reaction carousel. Rotate the carousel manually until the empty position exposes the cuvette wiper assembly at the rear of the water bath. Remove the two (2) screws [2] securing the cuvette wiper assembly. [1] Move the mounting plate [1] to the left and lift the plate out of the water bath. Remove the cuvette wiper mounting clip [3] and the cuvette wipers. Replacement Action VIDEO Install Cuvette Wipers and Wiper Plate Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Cuvette Wipers and Wiper Plate 1. 2. 3. 4. 5. Insert the new cuvette wipers [2] into the plate. [1] Snap the cuvette wiper mounting clip [3] into place on the mounting plate. [1] Reinstall the new wiper assembly into the water bath and secure it using two (2) screws. Manually rotate the carousel and verify that a cuvette segment passes centered between the wipers while the wipers contact both sides of the cuvette. Reinstall the cuvette segment. Prepare for Operation 1. 2. Replace the rear right cover and the rear center right cover. Power ON the processing module. Verification Procedures Order VP Description 1 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.03 Cup, ICT-Ref with Probes Version - 204738-101_4953_1 List/Part Numbers List/Part Number Description 7-205355-01 CUP, ICT REF, WITH PROBES Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.03 Cup, ICT-Ref with Probes Time Required 42 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Caution Step Caution Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. (Use the main circuit breaker located at the rear of the processing module.) Remove the covers. a. Front right cover [C] b. Center right cover [E] c. Rear right cover [K] Remove the cuvette segments located directly in front of the ICT reference solution cup. 2. 3. Remove ICT Preamp Module 1. 2. 3. Disconnect the cable from the LUI board [1]. Remove the two (2) screws [2] securing the LUI keypad to frame of instrument. Remove the screw [3] securing the ICT Preamp module and lift module out of instrument to access the ICT IREF Level Detect Board. Reference Disconnect CNN2 from ICT IREF Level Detect Board 1. Loosen the screw [4] securing the ICT IREF level detect board cover to the ICT preamp module and remove cover. Disconnect CNN2 from ICT IREF Level Detect Board (continued) 1. Disconnect ILS-2 [1] from CN2 on the ICT IREF level detect board. Remove wire ties from harness for removal of ICT reference solution cup wires. 2. Remove ICT Ref Cup 1. 2. 3. 4. 5. Remove the two (2) screws securing the ICT reference solution cup. Disconnect tubing number 12 [1] from the warming ring and tubing number 13 [2] from the ICT reference solution cup. Lift the ICT reference solution cup up to gain access to the tubing connection [3] between the warming ring and the cup. Disconnect the warming ring from the cup. Note Ensure that all wire ties are removed to allow removal of the wires for the ICT reference solution cup and its cable. Remove the ICT reference solution cup from the processing module. Replacement Action Steps Install ICT Ref Cup 1. 2. 3. 4. Connect the output side of the ICT warming ring [3] to the side of the ICT reference solution cup with the small tubing. Connect tubing number 12 to the warming ring inlet. [1] Connect tubing number 13 [2] to the flexible tubing coming out of the top of the ICT reference solution cup. Note Ensure that tubing number 13 is inserted at least 5 mm into the flexible tubing coming out of the top of the ICT reference solution cup. Install the ICT reference solution cup into Reference 5. Connect CNN2 from 1. ICT IREF Level Detect Board 2. 3. Install ICT Preamp Module 1. 2. 3. Prepare for Operation 1. 2. 3. 4. 5. 6. position and secure using two (2) screws. Secure cable from ICT reference solution cup with cable ties. Connect ILS-2 to CN2 [1] on the ICT IREF level detect board. Install the ICT IREF level detect board cover on the ICT preamp module. Secure the cable using wire ties. Slide the ICT Preamp module into the instrument and secure with screw [3]. Note Ensure cables are not stretched, pinched or kinked. Install the LUI keypad and secure with the two (2) screws [2]. Connect the LUI board cable [1]. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform System/Maintenance/As Needed/2131 Flush ICT Cup. Check for leaks. Install the cuvette segment removed from in front of the ICT reference solution cup. Install the covers: a. Front right cover [C] b. Center right cover [E] c. Rear right cover [K] Verification Procedures Order VP Description VP Detail / Note 2 6063 - Flush ICT Module Check for leaks. 1 P-618 ICT Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If Expected Results the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.04 Detector, Temp., Water Bath Version - 204738-101_4954_1 List/Part Numbers List/Part Number Description 7-205352-01 TEMP DETECTOR, WATER BATH, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.04 Detector, Temp., Water Bath Time Required 30 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 2. Drain the water bath. (System/Diagnostic/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the covers: d. Center left cover [D] e. Rear center left cover [I] f. Rear center right cover [J] g. Rear right cover [K] Remove the cuvette segment located in front of the water bath float switch and place in a safe location. 3. 4. 5. Remove Temperature Detector, Water Bath 1. 2. Disconnect the BTEM plug from the connector plate and remove the connector from the plate. Remove the screw securing the detector to the water bath. Note Cut the cable ties while removing the detector. Replacement Action Steps Install Temperature Detector, Water Bath 1. Prepare for Operation 1. 2. 2. 3. 4. Reference Install the detector into the water bath and secure using the screw. Insert the connector BTEM into the connector plate and plug in plug BTEM. Note Secure the cable with cable ties as it was prior to removal. Install the cuvette segment. For information on the startup procedure, refer to the ARCHITECT Reinstall the covers: System Operations Manual, Section 5: Operating instructions, a. Center left cover [D] Subsection: System startup, pause, and shutdown. b. Rear center left cover [I] c. Rear center right cover [J] d. Rear right cover [K] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 P-605 Water Bath Temperature Detector Adjustment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.05 Switch, Float, Water Bath Version - 204738-101_4955_1 List/Part Numbers List/Part Number Description 7-205354-01 FLOAT SWITCH, WATER BATH, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.05 Switch, Float, Water Bath Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 2. Drain the water bath. (System/Diagnostic/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the covers: e. Center left cover [D] f. Rear center left cover [I] g. Rear center right cover [J] h. Rear right cover [K] Remove the cuvette segment located in front of the water bath float switch and place in a safe location. 3. 4. 5. Remove Float Switch, Water Bath 1. Disconnect the float switch connector. [1] Remove Float Switch, Water Bath (continued) 1. 2. Remove the two screws [2] securing the float switch mounting bracket to the water bath and lift the mounting bracket out of the water bath. Unscrew the nut [3] securing the float switch to the mounting bracket and thread. Note Cut the cable ties while removing the detector. Replacement Action Steps Install Float Switch, Water Bath 1. 2. 3. Insert the float switch into the mounting bracket and secure using the nut. [3] Install the float switch mounting bracket into position on the water bath and secure using two (2) screws. [2] Connect the float switch connector. [1] Note Secure the cable with cable ties as Reference it was prior to removal. Prepare for Operation 1. 2. 3. 4. Install the cuvette segment. For information on the startup procedure, refer to the ARCHITECT Reinstall the covers: System Operations Manual, Section 5: Operating instructions, a. Center left cover [D] Subsection: System startup, pause, and shutdown. b. Rear center left cover [I] c. Rear center right cover [J] d. Rear right cover [K] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.06 Lens, Water Bath, Inner Version - 204738-101_4956_1 List/Part Numbers List/Part Number Description 2-89215-02 Gasket, Inner, Water Bath Lens Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.06 Lens, Water Bath, Inner Time Required 55 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Hot Surface Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath. (System/Diagnostic/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover Remove the rear right deck cover [K]. Remove the Lamp housing cover [G]. Remove the cuvette segments from the reaction carousel and place in a safe location. 2. 3. 4. 5. 6. Remove Source Lamp 1. 2. Remove the thumbscrew [1] securing the source lamp holder and remove the lamp holder. Lay the holder out of the way. Note There is no need to disconnect lamp from terminal block. Remove Optics Assembly 1. 2. 3. Remove Optics Assembly (continued) 1. Follow the cables from the Fan and Pre-Amp board and disconnect them (both, the shutter cable SHUT-SHAT and fan PHOTO-FAN). Remove the Pre-Amp cover [2] by loosening the left screw and removing the right screw with the ground cable. [1] Disconnect the OPT cable from the Pre-Amp board. While supporting the optics assembly from the bottom, remove the four (4) screws [3] securing the optics and remove it from the instrument. Remove Water Bath Lens 1. 2. 3. 4. Unscrew the lens holder. [1] Remove the outer gasket [2] Remove the lens [3] Remove the inner gasket. [4] Replacement Action Steps Install Water Bath Lens 1. 2. 3. Install Optics Assembly 1. 2. 3. 4. 5. 6. Install the inner gasket [4] on the inner side of the lens. Install the lens [3] and gaskets in place using the lens screw. [1] Install the outer gasket [2] on the outside of the lens. While supporting the optics assembly from the bottom, secure it to the processing module frame using four (4) screws [3]. Connect the pre-amp cable. [4] Install the Pre-Amp cover and ground wires, route the cables between the gasket material (of the cover) and the housing Insert the ground cable and secure the cover with the screws. Note Ensure that the cover slides into the guides in the bottom of the optics housing. The fit between the guides in the optic housing and the pre-amp cable cover is tight. Connect the shutter cable (SHUT to SHAT) Connect the fan (PHOTO-FAN) cable. Reference Install Source Lamp 1. Install source lamp holder in optics assembly and tighten the screw. Prepare for Operation 1. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2134 Change Water Bath. Check for leaks. Install the cuvette segments. Install the lamp housing cover [G]. Install the rear right deck cover [K]. Install right rear cover. 2. 3. 4. 5. 6. 7. 8. 9. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 1001 - Optics Trigger Sensor Check 2 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.07 Lens, Water Bath, Outer Version - 204738-101_4957_1 List/Part Numbers List/Part Number Description 2-89216-02 Gasket, outer, Water Bath Lens Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.07 Lens, Water Bath, Outer Time Required 55 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Hot Surface Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath. (System/Diagnostic/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover Remove the rear right deck cover [K]. Remove the Lamp housing cover [G]. Remove the cuvette segments from the reaction carousel and place in a safe location. 2. 3. 4. 5. 6. Remove Source Lamp 1. 2. Remove the thumbscrew [1] securing the source lamp holder and remove the lamp holder. Lay the holder out of the way. Note There is no need to disconnect lamp from terminal block. Remove Optics Assembly 1. 2. 3. 4. Follow the cables from the Fan and Pre-Amp board and disconnect them (both, the shutter cable SHUT-SHAT and fan PHOTO-FAN). Remove the Pre-Amp cover [2] by loosening the left screw and removing the right screw with the ground cable. [1] Disconnect the OPT cable from the Pre-Amp board. While supporting the optics assembly from the bottom, remove the four (4) screws [3] securing the optics and remove it from the instrument. Remove Water Bath Lens 1. 2. 3. 4. Unscrew the lens holder. [1] Remove the outer gasket [2] Remove the lens [3] Remove the inner gasket. [4] Replacement Action Steps Install Water Bath Lens 1. 2. 3. Install Optics Assembly 1. 2. 3. 4. 5. 6. Install the inner gasket on the inner side of the lens. Install the lens and gaskets in place using the lens screw. Install the outer gasket on the outside of the lens. While supporting the optics assembly from the bottom, secure it to the processing module frame using four (4) screws. Connect the pre-amp cable. Install the Pre-Amp cover and ground wires, route the cables between the gasket material (of the cover) and the housing. Insert the ground cable and secure the cover with the screws. Note Ensure that the cover slides into the guides in the bottom of the optics housing. The fit between the guides in the optic housing and the pre-amp cable cover is tight. Connect the shutter cable (SHUT to SHAT). Connect the fan (PHOTO-FAN) cable. Install Source Lamp 1. Install source lamp holder in optics assembly and tighten the screw. Prepare for Operation 1. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. 2. 3. Reference For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 4. 5. 6. 7. 8. 9. Perform 2134 Change Water Bath. Check for leaks. Install the cuvette segments. Install the lamp housing cover [G]. Install the rear right deck cover [K]. Install right rear cover. Verification Procedures Order VP Description 1 1001 - Optics Trigger Sensor Check 2 1008 - Optics Total Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N1.08 Warming Ring, Water Bath, C4 Version - 204738-101_4958_1 List/Part Numbers List/Part Number Description 7-205351-01 WARMING RING, WATER BATH, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. N1.08 Warming Ring, Water Bath, C4 Time Required 42 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath. (System/Diagnostic/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the covers: a. Front right cover [C] b. Center right cover [E] c. Lamp house cover [G] d. Rear center left cover [I] e. Rear center right cover [J] f. Rear right cover [K] Remove the cuvette segments from the reaction carousel and place. 2. 3. 4. Access Warming Ring 1. 2. 3. 4. 5. Remove the two (2) screws securing the ICT reference solution cup. Disconnect tubing number 12 [1] from the warming ring and tubing number 13 [2] from the ICT reference solution cup. Lift the ICT reference solution cup up to gain access to the tubing connection [3] between the warming ring and the cup. Disconnect the warming ring from the cup. Note Ensure that all wire ties are removed to allow removal of the wires for the ICT reference solution cup and its cable. Remove the ICT reference solution cup from the processing module. Remove Warming Ring 1. 2. 3. 4. 5. Remove the two (2) screws securing the cuvette wiper assembly [1] and remove the assembly. Remove the screw securing the temperature sensor [2] to the water bath and remove the sensor. Remove the screw securing the inlet water cover plate to the front of the water bath and remove the plate. [4] Remove the screw securing the inlet water cover plate to the rear of the water bath and remove the plate. [3] Lift the warming ring out of the water bath. Replacement Action Steps Install Warming Ring 1. 2. 3. 4. Install ICT Ref Cup 1. 2. 3. Install the warming ring into the water bath. Install the inlet water cover plates front [4] and rear [3] inside the water bath over the inlet port and secure using two screws. Install the temperature sensor [2] in the water bath using two screws. Install the cuvette wiper assembly [1] in the water bath and secure using two screws. Connect the output side of the ICT warming ring [3] to the side of the ICT reference solution cup with the small tubing. Connect tubing number 12 to the warming ring inlet. [1] Connect tubing number 13 [2] to the flexible tubing coming out of the top of the ICT reference solution cup. Note Ensure that tubing number 13 is inserted at least 5 mm into the flexible tubing coming out of the Reference 4. 5. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. top of the ICT reference solution cup. Install the ICT reference solution cup into position and secure using two screws. Secure cable from ICT reference solution cup with cable ties. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2131 Flush ICT Cup. Check for leaks. Install the cuvette segments. Reinstall the covers: a. Front right cover [C] b. Center right cover [E] c. Lamp house cover [G] d. Rear center left cover [I] e. Rear center right cover [J] f. Rear right cover [K] Verification Procedures Order VP Description VP Detail / Note 2 6063 - Flush ICT Module Check for leaks. 1 P-618 ICT Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N2.01 Incubator, Water Bath Heater Version - 204738-101_4959_1 List/Part Numbers List/Part Number Description 7-205356-01 INCUBATOR, WTR BTH HTR Inspect tools for damage, ensure calibration is not expired and replace if necessary. N2.01 Incubator, Water Bath Heater Time Required 45 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 2. Drain the water bath. (System/Diagnostic/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the left rear cover: 3. 4. Remove Heater Assembly Note 1. 2. 3. 4. 5. You need a container for the water from the heater assembly. Access the service drain, located in the heater assembly and drain it. Replace the plug in the service drain (and tighten tubing clamp) after all of the water has been drained from the heater assembly. Remove the top two (2) screws securing the heater control cover and remove it. From the controller board disconnect: • TBC • TBC2 • TBC1 Remove the three screws securing the TBC board support mounting bracket [4] to the processing module frame to gain access to the water bath heather. Remove Heater Assembly (continued) 1. 2. 3. 4. 5. From the heater bottle, disconnect the tubing [1] closest to the rear of the module. Loosen the tubing clamp and disconnect the tubing [2] from the elbow at the water pump. Loosen the tubing clamp and disconnect tubing connected to the 3-way fitting [3]. Remove the two (2) screws from the base of the heater assembly and slide the assembly toward the rear of the processing module to allow access to the tubing at the elbow at the water pump. Remove the water bath heater from the processing module. Replacement Action Steps Install Heater Assembly 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. 5. 6. Reference Install the water bath heather into the processing module and connect the tubing [2] to the elbow at the water pump. Connect tubing to heater bottle [1] and 3-way fitting [3]. Note Ensure the tubing clamps are tight. Slide the heater assembly into the processing module and secure it to the processing module frame using two screws. Tighten the three (3) screws securing the TBC board support mounting bracket [4] to the processing module frame. In controller board connect: a. TBC b. TBC2 c. TBC1 Note Do not install the heater control cover at this time. Access is required during the temperature adjustment For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2134 Change Water Bath. Check for leaks. Perform P-619 Water Bath Heater Adjustment. 7. 8. Install the heater control cover using two screws. Install the rear left cover. Verification Procedures Order VP Description 1 P-605 Water Bath Temperature Detector Adjustment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. N2.02 Bd., Temp. Bath Control, Heater Version - 204738-101_4960_1 List/Part Numbers List/Part Number Description 7-205360-01 BD, TBC 7-205360-02 Bd, TBC New Style Inspect tools for damage, ensure calibration is not expired and replace if necessary. N2.02 Bd., Temp. Bath Control, Heater Time Required 15 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite 1. The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover. 2. Remove TBC Heater Board 1. 2. Remove the two (2) screws securing the heater control cover and disconnect all cables from the TBC board. Note Mark the unlabeled cables. Remove the board from the standoffs. Reference Replacement Action Steps Install TBC Heater Board 1. 2. Install the board onto the standoffs. Connect all cables to the TBC board. Note Do not install the heater control cover at this time. Access is required during the temperature adjustment procedure. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. 7. 8. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 2134 Change Water Bath. Check for leaks. Perform P-619 Water Bath Heater Adjustment. Install the heater control cover using two (2) screws. Install the rear left cover Verification Procedures Order VP Description 1 P-605 Water Bath Temperature Detector Adjustment N/A G110 - After repair is complete, verify per VP Detail / Note Expected Results released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Q2.02 Monitor, Touchscreen Version - 96728-111_220_1 List/Part Numbers List/Part Number Description 07D03-17 Monitor 17" Touchscreen Inspect tools for damage, ensure calibration is not expired and replace if necessary. Q2.02 Monitor, Touchscreen Time Required Tools/Materials 20 min None WARNING Potential Biohazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped, or Offline status. Perform SCC Shutdown 1. Access the Snapshot screen by selecting Overview and then Snapshot. 2. Select F3-Shutdown. 3. Select OK to confirm the Shutdown request. 4. When the screen turns blue, and a message about shutting down the SCC displays, press the Ctrl+Alt+Delete keys simultaneously. 5. A popup message displays. Select Shutdown, then OK. Note Wait while the software applications shut down. A message displays informing you that it is safe to turn off the computer. 6. Press the power switch on the front of the central processing unit (CPU) to power OFF the system control center. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. For information on the shutdown procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Remove Monitor 1. Press the power switch on the front of the monitor to power OFF the monitor. 2. Unplug the monitor power cord from the wall outlet or UPS. 3. Unscrew and disconnect the monitor cable from the port on the back of the CPU. [1] 4. Unscrew and disconnect the touchscreen cable from the port on the back of the CPU. [2] Replacement Action Install Monitor Steps Reference 1. Connect the monitor cable to the port on the back of the CPU. Secure the cable to the CPU by screwing in the connector. [1] 2. Connect the touchscreen cable to the port on the back of the CPU. Secure the cable to the CPU by screwing in the connector. [2] 3. Connect the monitor power cord to the monitor. [3] 4. Plug the power cord into the wall outlet or UPS. Prepare for Operation 1. Power OFF the processing module(s). Note For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. The power to the processing module(s) must be OFF prior to turning on the power to the system control center. If the processing module(s) power is ON, communication is not properly initialized between the system components. 2. Power ON the system control center by pressing the power switch on the front of the CPU. 3. Power ON the monitor. Note Do not go to the next step until the system control center displays the Snapshot screen. 4. Power ON the processing module(s). 5. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G11 - Monitor Verify Monitor turns on and displays the intended screen. 2 P-205 Calibrate Touchscreen N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R1.01 Mixer Assembly Version - 204738-101_4961_1 List/Part Numbers List/Part Number Description 7-205197-01 MIXER ASSY., C4 7-205197-02 MIXER ASSY., C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R1.01 Mixer Assembly Time Required 45 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover. Remove the left rear cover. Remove the rear right deck cover [K]. Remove the rear center left deck cover [I]. Remove the rear center right deck cover [J]. Remove the mixers. 2. 3. 4. 5. 6. 7. Remove Mixer Assembly 1. 2. 3. 4. Remove the two (2) screws securing the mixer cover and remove the cover. Disconnect the following cables from the bottom of the mixer assembly: • ST1AM • ST2AM • STRD • STRM • STRE-S • STR-S Remove the four (4) screws [1] from the base of the mixer assembly. Lower the mixer assembly slightly to clear the positioning pins from the frame and remove the mixer assembly from the processing module frame. Note Do not lift with up/down mechanism, this can damage the cam follower spring. Replacement Action Steps Install Mixer Assembly 1. 2. 3. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. 8. 9. Reference Install the mixer assembly onto the processing module frame and secure using four (4) screws [1]. Connect the following cables to the mixer assembly: • ST1AM • ST2AM • STRD • STRM • STRE-S • STR-S Install the mixer cover using two (2) screws. Install mixers. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform P-620 Mixer Alignment. Install the rear center right deck cover [J]. Install the rear center left deck cover [I]. Install the rear right deck cover [K]. Install the left rear cover. Install the right rear cover. Verification Procedures Order VP Description 1 3126 - Mixer Vibration Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R1.02 Mixer Elevator Assembly Version - 204738-101_4962_1 List/Part Numbers List/Part Number Description 7-205376-01 MIXER, UP DWN MECH, C4 7-205376-02 MIXER, UP DWN MECH, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R1.02 Mixer Elevator Assembly Time Required 45 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover. Remove the left rear cover. Remove the rear right deck cover [K]. Remove the rear center left deck cover [I]. Remove the rear center right deck cover [J]. Remove the mixers. 2. 3. 4. 5. 6. 7. Remove Mixer Head 1. Assembly 2. 3. 4. Remove the two (2) screws securing the mixer cover and remove the cover. Disconnect the following cables from the mixer assembly connector mount: • ST1AM • ST2AM • STRD • STRM • STRE-S • STR-S Remove the following jacks from the connector mount [1] on the mixer assembly: • STRA • STRB • STRD • STR-S Cut the wire ties [3] & [4] securing the wire cables to the mixer assembly. 5. 6. Remove Mixer Elevator Assembly 1. 2. Remove the three (3) screws [2] from the mounting plate on the mixer head assembly. Lift the mixer head straight up off of the mixer elevator. Remove the four (4) screws [1] from the base of the mixer elevator assembly. Lower the mixer assembly slightly to clear the positioning pins from the frame and remove the mixer elevator assembly from the processing module frame. Note Do not lift with up/down mechanism, this can damage the cam follower spring. Replacement Action Steps Install Mixer Elevator Assembly 1. Install the mixer elevator onto the processing module frame and secure using four (4) screws. [1] Reference Install Mixer Head Assembly 1. 2. 3. 4. 5. Install the mixer head onto the guide pins on the mixer elevator and secure using three (3) screws. [2] Install the jacks to the connector mount [1] on the mixer assembly: • STRA • STRB • STRD • STR-S Connect the following plugs to the mixer assembly: • ST1AM • ST2AM • STRD • STRM • STRE-S • STR-S Secure the mixer wire cables to the mixer assembly with wire ties [3] & [4]. Install the mixer cover using two (2) screws. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. 8. 9. Install mixers. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform P-620 Mixer Alignment. Install the rear center right deck cover [J]. Install the rear center left deck cover [I]. Install the rear right deck cover [K]. Install the left rear cover. Install the right rear cover. Verification Procedures Order VP Description 1 3126 - Mixer Vibration Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R2.01 Pipettor, Sample Complete Version - 204738-101_4963_1 List/Part Numbers List/Part Number Description 7-205171-01 PIPETTOR, SAMPLE, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R2.01 Pipettor, Sample Complete Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the following deck covers: • Center left cover [D] • Front center cover [B] • Front right cover [C] • Center right cover [E] Remove the Sample Pipettor head cover Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Disconnect the tubing from the sample probe. Remove the probe from the pipettor. 2. 3. 4. 5. 6. SCRUSH and STLLS bracket 1. 2. 3. 4. Loosen the two (2) screws [1] that hold the bracket for the SCRUSH and STLLS connectors located by the LUI keyboard. Lift the SCRUSH and STLLS connector board and disconnect the following cables [2]: • SCRUSH • STLLS Unscrew the ground cables [3], remove only the one that goes into the flexible cable duct (put back the screw to hold the ground cables attached to the bracket). Remove the flexible cable sleeve from the bracket. [4] Remove Pipettor, Sample Assembly 1. 2. 3. Remove the four screws securing the Sample Pipettor. [1] Disconnect and secure the following cables from the pipettor connector plate: [2] • AAS • SAAM • SAEM Carefully lift the Sample Pipettor assembly to prevent damage to the cables or pinching the sample tubing. Remove Pipettor, Sample Assembly (continued) 1. 2. 3. Separate tubing from the metal fitting. [3] Secure tubing to reconnect the after replacement. Remove the Sample Pipettor assembly from the instrument. Replacement Action Steps Install Pipettor, Sample Assembly 1. 2. 3. 4. Connect and mount SCRUSH and STLLS bracket. 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Reference Connect the tubing fully onto the metal fitting. Carefully place the Sample Pipettor arm in position and guide the flexible cable sleeve towards the LUI board to avoid damaging the cable and wires. Connect the following cables: • SAAS • SAAM • SAEM Secure the Sample Pipettor with the four screws. Insert the flexible cable sleeve into the clamp. Install the ground cable and secure both ground cables with the screw. Connect the following cables: • SCRUSH • STLLS Install the SCRUSH and STLLS bracket and secure using the two (2) screws. Install the probe. For information on the startup procedure, refer to the ARCHITECT Connect the tubing to the sample probe. System Operations Manual, Section 5: Operating instructions, Tighten the screw for the probe ground wire. Subsection: System startup, pause, and shutdown. Install the Sample Pipettor head cover Replace the following deck covers: • Center left cover [D] • Front center cover [B] • Front right cover [C] • Center right cover [E] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 1 1101 - Robotics Test Tool 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R2.02 Tubing, Sample Pipettor, Inner Version - 204738-101_4964_1 List/Part Numbers List/Part Number Description 7-202578-01 Tubing, Samp Pippetor, inner 7-93238-01 Tubing, Sample Pipettor, Inner Inspect tools for damage, ensure calibration is not expired and replace if necessary. R2.02 Tubing, Sample Pipettor, Inner Time Required 40 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 2. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove: a. Center Deck-Center left cover (D) b. Front Deck-Front center cover (B) c. Front Deck-Front right cover (C) d. Center Deck-center right cover (E) Remove the Sample Pipettor head cover. Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Disconnect the tubing from the sample probe. Remove the probe from the pipettor. 3. 4. 5. 6. 7. SCRUSH and STLLS bracket 1. 2. 3. 4. Loosen the two (2) screws [1] that hold the bracket for the SCRUSH and STLLS connectors located by the LUI keyboard. Release and lift the SCRUSH and STLLS connector board, disconnect the following cables: [2] a. SCRUSH b. STLLS Unscrew the ground cables [3], remove only the one that goes into the flexible cable duct. Note Put back the screw to hold the ground cables attached to the bracket. Release the flexible cable duct from the bracket. [4] Remove Pipettor, Sample Assembly 1. 2. 3. Remove the four (4) screws securing the Sample Pipettor. [1] Disconnect the following cables from the pipettor and secure them: [2] a. SAAS b. SAAM c. SAEM Carefully lift the Sample Pipettor assembly to prevent any damage to the cables or pinching the sample tubing. Remove Pipettor, Sample Assembly (continued) 1. 2. Remove Tubing, Sample Arm, Inner 1. 2. 3. 4. 5. 6. 7. Replacement Separate the tubing from the connector. Secure the tubing to be able to reconnect the tubing after being replaced. [3] Remove the Sample Pipettor assembly from the instrument. Remove the protecting sleeve and identification clamps from the tubing. Loosen the bracket screw closer to the head of the Sample Pipettor Assembly. Remove the screw from the bottom part of the bracket (close to the motor) to gain access to the plastic cover and the flexible metal sleeve. Remove the screw from the plastic cover. [1] Release the tabs to open the cover and stretch the flexible metal sleeve. [2] Note the position for the cable ducts. Pull the tubing to remove it from the pipettor. Action Steps Install Tubing, Sample Arm, Inner 1. 2. 3. 4. 5. 6. Install Pipettor, Sample Assembly 1. 2. 3. 4. Connect and mount SCRUSH and STLLS bracket. 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. 5. 6. Reference Insert the new tubing from the Sample Pipettor head toward the bottom. [1] Feed the tubing (without the protective sleeve) through the flexible metal sleeve and through the flexible cable duct. Place the protective sleeve and identification clamps on the tubing. Place back the cable ducts and snap the plastic cover. Secure the plastic cover with the screw, make sure the tabs are holding the plastic cover in place. Place back the metal bracket, insert the screw and tighten both ends. Connect the tubing to the connector assuring the connector will go all the way into the tubing. Carefully insert the Sample Pipettor arm in position. Note Make sure you are guiding the flexible cable duct towards the LUI board. Connect the cables: a. SAAS b. SAAM c. SAEM Secure the Sample Pipettor with the four (4) screws. Guide the flexible cable cover with its cables and tubing to the LUI section. Snap the cable cover back into the clamp. Install the ground cable and secure both ground cables with the screw. Connect the following cables: a. SCRUSH b. STLLS Place back the SCRUSH and STLLS bracket. Tighten the two (2) screws that hold the bracket. Install the probe. For information on the startup procedure, refer to the ARCHITECT Connect the tubing to the sample probe. System Operations Manual, Section 5: Operating instructions, Tighten the screw for the probe ground wire. Subsection: System startup, pause, and shutdown. Install the Sample Pipettor head cover. Replace: a. Center Deck-Center left cover (D) b. Front Deck-Front center cover (B) c. Front Deck-Front right cover (C) d. Center Deck-center right cover (E) e. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 1 1101 - Robotics Test Tool 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R2.03 Photo Sensor, Pitch, Sample Pipettor Version - 204738-101_4965_1 List/Part Numbers List/Part Number Description 7-204135-01 Sensor, pitch, pipettor, new style Inspect tools for damage, ensure calibration is not expired and replace if necessary. R2.03 Photo Sensor, Pitch, Sample Pipettor Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or For further information, refer to the ARCHITECT System Operations Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the following deck covers: a. Center left cover [D] b. Front center cover [B] Remove the Sample Pipettor head cover. Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Disconnect the tubing from the sample probe. Remove the probe from the pipettor. 2. 3. 4. 5. 6. Remove Sensor, Pitch, Pipettor 1. 2. 3. Disconnect the ROT-P cable from the sensor. [1] Remove the two (2) screws securing the Pitch sensor bracket to the Sample Pipettor assembly. [2] Remove the screw that holds the Pitch sensor to the bracket. [3] Reference Replacement Action Steps Install Sensor, Pitch, Pipettor 1. 2. 3. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. 2. 3. 4. 5. Reference Mount the Pitch sensor to the bracket. Secure the Pitch sensor bracket to the Sample Pipettor assembly. Connect the ROT-P cable from the sensor. Install the probe. Connect the tubing to the sample probe. Tighten the screw for the probe ground wire. Install the Sample Pipettor head cover. Replace the following deck covers: a. Center left cover [D] b. Front center cover [B] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Install the probe. For information on the startup procedure, refer to the ARCHITECT Connect the tubing to the sample probe. System Operations Manual, Section 5: Operating instructions, Tighten the screw for the probe ground wire. Subsection: System startup, pause, and shutdown. Install the Sample Pipettor head cover. Replace: a. Center Deck-Center left cover (D) b. Front Deck-Front center cover (B) c. Front Deck-Front right cover (C) d. Center Deck-center right cover (E) e. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Note To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 1120 - Sample Pipettor Calibration N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R3.01 Bd., LLS Pre-Amp, Sample Pipettor Version - 204738-101_4966_1 List/Part Numbers List/Part Number Description 7-202579-01 Bd., LLS Pre-Amp, S Pipettor 16 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R3.01 Bd., LLS Pre-Amp, Sample Pipettor Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Removal Action Steps Prerequisite 1. The processing module must be in Stopped or For further information, refer to the ARCHITECT System Operations Offline status. Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the Sample Pipettor head cover. [1] Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. [2] Disconnect the tubing from the sample probe. Remove the probe from the pipettor. [3] Remove the tubing holder (board cover). 2. 3. 4. 5. 6. Remove LLS PreAmp board 1. 2. 3. Disconnect the SLVL cable [1] from the LLS Pre-Amp board. Move the ground shield bracket [2] to remove the screw [3] that secures the board. Remove the LLS Pre-Amp board. Reference Replacement Action Steps Install the LLS PreAmp board 1. 2. 3. 4. Prepare for Operation Reference Install the LLS Pre-Amp board. Note When installing the board, ensure the sensor is under the flag. Secure the board using the one (1) screw. Connect the SLVL cable. Place the tubing holder into position. 1. 2. 3. Install the probe. For information on the startup procedure, refer to the ARCHITECT Connect the tubing to the sample probe. System Operations Manual, Section 5: Operating instructions, Place the probe ground wires under the Subsection: System startup, pause, and shutdown. ground screws and tighten the screws. 4. Install the Sample Pipettor head cover. 5. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 1120 - Sample Pipettor Calibration 2 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R3.02 Probe Slide Mechanism Version - 204738-101_4967_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. R3.02 Probe Slide Mechanism Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the Sample Pipettor head cover. [1] Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. [2] Disconnect the tubing from the sample probe. Remove the probe from the pipettor. [3] Remove the tubing holder (board cover). 2. 3. 4. 5. 6. Remove Pipettor 1. LLS Pre-Amp board 2. 3. Disconnect the SLVL cable [1] from the LLS Pre-Amp board. Rotate the ground shield bracket [2] clockwise (CW) to remove the screw [3] securing the board. Remove the LLS Pre-Amp board. Remove Probe Slide Mechanism 1. 2. Loosen the two (2) screws securing the Probe Slide Mechanism. [1] Carefully remove the Probe Slide mechanism. Note There are two "T" shaped metal shims between the pipettor head and the slide mechanism, retain the shims to use during installation of the new mechanism. Replacement Action Steps Remove Probe Slide Mechanism 1. 2. Insert the Probe Slide mechanism. Note Ensure the two (2) "T" shaped metal shims are installed between the probe slide mechanism and the pipettor head. Secure the Probe slide mechanism using the two (2) screws. [1] Note Make sure the sliding mechanism sits evenly and moves smoothly. Reference Install the Pipettor 1. LLS Pre-Amp board 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. Install the LLS Pre-Amp board. Note When installing the board, ensure the sensor is under the flag and firmly seated. Secure the board using the one (1) screw. Connect the SLVL cable. Place the tubing holder into position. Install the probe. Connect the tubing to the sample probe. Secure the ground wire with one (1) screw to the probe mount. Install the Sample Pipettor head cover. Power ON the processing module. Note Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 1120 - Sample Pipettor Calibration 2 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R4.01 Pipettor, Reagent R1 Assembly Version - 204738-101_4968_1 List/Part Numbers List/Part Number Description 7-205172-01 PIPETTOR, REAGENT R1, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R4.01 Pipettor, Reagent R1 Assembly Time Required 35 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the Reagent pipettor head cover. Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Remove the probe from the pipettor. Note The end of the reagent probe tubing attached to the probe has been tapered and removal from the probe could damage the fit between the probe and tubing. Remove the following deck covers: a. Center left cover [D] b. Rear left cover [H] c. Rear center left cover [I] 2. 3. 4. 5. Remove Pipettor, Reagent Assembly 1. 2. 3. 4. Locate the R1 tubing and separate the tubing from the metal fitting and secure the tubing. [2] Disconnect the following cables from the pipettor connector plate: [1] • R1SENS • R1ASOL • 1CRASH • R1LLS • R1AA • R1AE Remove the three (3) screws securing the Reagent pipettor. [3] Lift the Reagent pipettor assembly carefully to prevent any damage to the cables or pinching the tubing. Replacement Action Steps Install Pipettor, Reagent Assembly 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Reference Insert the Reagent pipettor arm carefully in position, ensuring that the wires and tubing are not crushed or crimped. Secure the Reagent pipettor with the three (3) screws. Connect the following cables to the pipettor connector plate: • R1SENS • R1ASOL • R1CRASH • R1LLS • R1AA • R1AE Connect the R1 tubing fully onto the metal fitting. Install the probe. Connect the tubing to the reagent probe. Tighten the screw for the probe ground wire. Install the Reagent pipettor head cover. Replace the following deck covers: a. Center left cover [D] b. Rear left cover [H] c. Rear center left cover [I] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 2 1121 - R1 Pipettor Calibration 1 1101 - Robotics Test Tool 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R4.02 Pipettor, Reagent R2 Assembly Version - 204738-101_4969_1 List/Part Numbers List/Part Number Description 7-205173-01 PIPETTOR, REAGENT R2, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R4.02 Pipettor, Reagent R2 Assembly Time Required 35 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the Reagent pipettor head cover Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Remove the probe from the pipettor. Remove the following deck covers: • Front left cover [A] • Center left cover [D] • Front center cover [B] • Rear left cover [H] • Rear center left cover [I] 2. 3. 4. 5. Remove Pipettor, Reagent Assembly 1. 2. 3. 4. Locate the R2 tubing and separate the tubing from the metal fitting and secure the tubing. [1] Disconnect the following cables from the pipettor: [2] • R2SENS • R2ASOL • R2CRASH • R2LLS • R2AA • R2AE Remove the three (3) screws securing the Reagent pipettor. [3] Lift the Reagent pipettor assembly, carefully to prevent any damage to the cables or pinching the tubing. Replacement Action Steps Install Pipettor, Reagent Assembly 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Reference Carefully insert the Reagent pipettor arm in position, ensuring that the wires and tubing are not crushed or crimped. Secure the Reagent pipettor with the three (3) screws. Connect the cables: • R2SENS • R2ASOL • R2CRASH • R2LLS • R2AA • R2AE Connect the R2 tubing fully onto the metal fitting. Install the probe. Connect the tubing to the reagent probe. Tighten the screw for the probe ground wire. Install the Reagent pipettor head cover. Replace the following deck covers: • Front left cover [A] • Center left cover (D) • Front center cover [B] • Rear left cover [H] • Rear center left cover [I] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 2 1122 - R2 Pipettor Calibration 1 1101 - Robotics Test Tool 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, VP Detail / Note Expected Results ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R4.03 Tubing, R1 Pipettor Inner Version - 204738-101_4970_1 List/Part Numbers List/Part Number Description 7-93246-01 Tbg, R1 pipettor inne Inspect tools for damage, ensure calibration is not expired and replace if necessary. R4.03 Tubing, R1 Pipettor Inner Time Required 50 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the Reagent pipettor head cover. Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Remove the probe from the pipettor. Remove the following deck covers: • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] 2. 3. 4. 5. Remove Pipettor, Reagent Assembly 1. 2. 3. 4. Locate the R1 tubing and separate the tubing from the metal fitting and secure the tubing. [2] Disconnect the following cables from the pipettor: [1] • R2SENS • R2ASOL • R2CRASH • R2LLS • R2AA • R2AE Remove the three (3) screws securing the Reagent pipettor. [3] Lift the Reagent pipettor assembly, carefully to prevent any damage to the cables or pinching the tubing. Remove Reagent Pipettor Inner Tubing 1. 2. 3. 4. 5. Remove the protecting sleeve and identification clamps from the tubing. Remove the metal bracket by removing the two (2) screws to gain access to the plastic cover and the flexible metal sleeve. [1] Remove the one (1) screw from the plastic cover. [2] Release the tabs to open the cover and stretch the flexible metal sleeve. [3] Note Notice the position for the cable ducts. Pull the tubing to remove it from the pipettor. Replacement Action Steps Install Tubing, Reagent Arm, Inner 1. 2. 3. 4. 5. Insert the new tubing from the Reagent Pipettor head towards the bottom. Feed the tubing (without the protective sleeve) through the flexible metal sleeve. Place the protective sleeve and identification clamps on the tubing. Install the cable ducts and snap the plastic cover. Secure the plastic cover with the one (1) screw, and make sure the tabs are holding Reference 6. Install Pipettor, Reagent Assembly 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. the plastic cover in place. Install the metal bracket, insert the two (2) screws and tighten both ends. Insert the Reagent pipettor arm, carefully in position, ensuring that the wires and tubing are not crushed or crimped. Secure the Reagent pipettor with the three (3) screws. Connect the cables to the pipettor connector plate: • R1SENS • R1ASOL • R1CRASH • R1LLS • R1AA • R1AE Connect the R1 tubing fully onto the metal fitting. Install the probe. Connect the tubing to the reagent probe. Tighten the screw for the probe ground wire. Install the Reagent pipettor head cover. Replace the following deck covers: • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 2 1121 - R1 Pipettor Calibration 1 1101 - Robotics Test Tool 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R4.04 Tubing, R2 Pipettor Inner Version - 204738-101_4971_1 List/Part Numbers List/Part Number Description 7-93247-01 Tbg, R2 pipettor inner Inspect tools for damage, ensure calibration is not expired and replace if necessary. R4.04 Tubing, R2 Pipettor Inner Time Required 50 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the reagent pipettor head cover Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Remove the probe from the pipettor. Remove the following deck covers: • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] 2. 3. 4. 5. Remove Pipettor, Reagent Assembly 1. 2. 3. 4. Locate the R2 tubing and separate the tubing from the metal fitting and secure the tubing. [1] Disconnect the following cables from the pipettor: [2] • R2SENS • R2ASOL • R2CRASH • R2LLS • R2AA • R2AE Remove the three (3) screws securing the Reagent pipettor. [3] Lift the Reagent pipettor assembly, carefully to prevent any damage to the cables or pinching the tubing. Remove Reagent Pipettor Inner Tubing 1. 2. 3. 4. 5. Remove the protecting sleeve and identification clamps from the tubing. Remove the metal bracket by removing the two (2) screws to gain access to the plastic cover and the flexible metal sleeve. [1] Remove the one (1) screw from the plastic cover. [2] Release the tabs to open the cover and stretch the flexible metal sleeve. [3] Note Notice the position for the cable ducts. Pull the tubing to remove it from the pipettor Replacement Action Steps Install Tubing, Reagent Arm, Inner 1. 2. 3. 4. 5. Insert the new tubing from the Reagent Pipettor head towards the bottom. Feed the tubing (without the protective sleeve) through the flexible metal sleeve. Place the protective sleeve and identification clamps on the tubing. Install the cable ducts and snap the plastic cover. Secure the plastic cover with the one (1) screw, and make sure the tabs are holding the plastic cover in place. Reference Install Pipettor, Reagent Assembly 6. Install the metal bracket, insert the two (2) screw and tighten both ends. 1. Insert the Reagent pipettor arm, carefully in position, ensuring that the wires and tubing are not crushed or crimped. Secure the Reagent pipettor with the three (3) screws. Connect the cables to the pipettor connector plate: • R2SENS • R2ASOL • R2CRASH • R2LLS • R2AA • R2AE Connect the R2 tubing fully onto the metal fitting. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. 6. 7. Install the probe. Connect the tubing to the reagent probe. Tighten the screw for the probe ground wire. Install the Reagent pipettor head cover. Replace the following deck covers: • Center left cover [D] • Rear left cover [H] • Rear center left cover [I] Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 2 1122 - R2 Pipettor Calibration 1 1101 - Robotics Test Tool 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R5.01 Spring, Crash Detect, Reagent Pipettors Version - 204738-101_4972_1 List/Part Numbers List/Part Number Description 2-89380-02 Spring,crash detect, all arms Inspect tools for damage, ensure calibration is not expired and replace if necessary. R5.01 Spring, Crash Detect, Reagent Pipettors Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the Reagent pipettor head cover. Loosen the screw for the probe ground wire and remove the probe ground wire from under the screw. Remove the probe from the pipettor. 2. 3. 4. Remove Spring, Crash Detect 1. 2. Remove the screw from the bottom of the spring. Remove the screw from the top of the spring. Replacement Action Steps Install Spring, Crash 1. Detect 2. Insert the screw [1] for the top of the spring with the new spring. Insert the screw [2] for the bottom of the spring. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. Install the probe. For information on the startup procedure, refer to the ARCHITECT Connect the tubing to the reagent probe. System Operations Manual, Section 5: Operating instructions, Tighten the screw for the probe ground wire. Subsection: System startup, pause, and shutdown. Install the Reagent pipettor head cover. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 1121 - R1 Pipettor Calibration 2 1122 - R2 Pipettor Calibration 3 5405 - Crash Sensor Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.01 Washer, Cuvette Version - 204738-101_4973_1 List/Part Numbers List/Part Number Description 7-205198-01 CUVETTE WASHER, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.01 Washer, Cuvette Time Required 25 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover. Turn the peristaltic pump head counterclockwise (CCW) about 1/16 of a turn. Pull upward on the pump head to remove the head. Remove the front right deck cover [C]. Remove the rear right deck cover [K]. Remove the center right deck cover [E]. 2. 3. 4. 5. 6. 7. Remove Cuvette Washer Assembly 1. 2. 3. Remove all tubing from nozzles #1 through #8. Loosen the one (1) thumbscrew securing the cuvette wash head to the cuvette wash elevator and remove the wash head. [1] Remove the two (2) screws securing the mounting bracket and remove the mounting bracket. [2] Remove Cuvette Washer Assembly (continued) 1. 2. 3. Disconnect plugs WE-S and WE-M from the cuvette wash elevator. Support the elevator while removing the six (6) hex screws that secure the elevator to the processing module upper mounting plate. Lower the elevator and remove to the rear of the module. Replacement Action Steps Install Cuvette Washer Assembly 1. 2. Install Cuvette Washer Assembly (continued) 1. 2. 3. Install the elevator into position on the underside of the mounting plate and secure with six (6) hex screws. Install the mounting bracket using two (2) screws [2]. Connect cables WE-S and WE-M. Install the cuvette wash head on the cuvette wash elevator and secure using one (1) thumbscrew. [1] Install tubing on nozzles #1 through #8. Reference Install Peristaltic Pump Head and Front Right Rear Deck Cover 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. Install the front right deck cover. [2] Align the notch with the peristaltic motor tang. Align the attachment locking tabs on the peristaltic pump head with the recesses on the pump. [1] Turn head clockwise (CW) about 1/16 of a turn until it snap locks in place. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform P-621 Cuvette Washer Alignment. Select the As Needed category from the For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 5. 6. 7. 8. 9. Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Install the center right deck cover [E]. Install the rear right deck cover [K]. Install the right rear cover. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.02 Sensor, Photo Multiple Use Version - 204738-101_4974_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.02 Sensor, Photo Multiple Use Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the right rear cover. 2. Remove Photo Sensor 1. 2. Disconnect the cable from sensor [1]. Remove the one (1) screw securing the sensor [2] and remove the sensor. Note The cuvette washer assembly may be removed by performing R6.01 Washer, Cuvette to have better access to the D-OK and the UP sensors. Replacement Action Steps Install Photo Sensor 1. 2. Install the sensor [2] using one (1) screw. Connect the cable [1] to the sensor. Reference Prepare for Operation 1. 2. 3. 4. 5. 6. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Perform P-621 Cuvette Washer Alignment. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Install the right rear cover. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.03 Nozzle, #1, #4, & #5 Version - 204738-101_4975_2 List/Part Numbers List/Part Number Description 7-205228-01 NOZZLE, C4, #1, #4, #5 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.03 Nozzle, #1, #4, & #5 Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Nozzle, #1, #4, or #5 1. Remove the incoming and waste tubing [1] from the nozzle. Remove the screw [2] securing the nozzle to the mounting plate. Lift the nozzle straight up.[3] 2. 3. Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note Video only shows removal of nozzle 1. However, same instructions apply to removal of nozzles 4 and 5. Replacement Action Steps Install Nozzle, #1, #4, or #5 1. 2. Reference Place the nozzle on the mounting plate [3] and secure with one (1) screw [2]. Install the incoming and waste tubing [1] on the nozzle. [3] (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note Video only shows replacement of nozzle 1. However, same instructions apply to replacement of nozzles 4 and 5. Prepare for Operation 1. To change the status of the processing module and For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and 2. 3. 4. sample handler from Stopped to Ready, perform the Startup procedure. Perform P-621 Cuvette Washer Alignment. Select System Maintenance, As Needed and perform 6052 Wash Cuvettes. Install the cuvette wash head cover. shutdown. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks Expected Results Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.04 #2 & #3, Wash Solution Version - 204738-101_4976_1 List/Part Numbers List/Part Number Description 7-205229-01 NOZZLE, C4, #2, #3 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.04 # 2 & #3, Wash Solution Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Nozzle, # 2 1. or 3, Wash Solution 2. Remove the incoming and waste tubing [1] from the nozzle. Remove the screw [2] securing the nozzle to the mounting plate. Lift the nozzle straight up. 3. Replacement Action Steps Reference Install Nozzle, # 2 or 3, Wash Solution 1. 2. 3. Prepare for Operation 1. Place the nozzle on the mounting plate and secure with one (1) screw [2]. Install the incoming and waste tubing on the nozzle. [1] Install the cuvette wash head cover. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.05 Nozzle, #6 Blank Version - 204738-101_4977_2 List/Part Numbers List/Part Number Description 7-205230-01 NOZZLE, C4, #6 BLANK 7-205231-01 NOZZLE, C4, #7 SUCTION Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.05 Nozzle, #6, Blank Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Nozzle, #6, Blank 1. Remove the incoming and waste tubing [1] from nozzle #6. Remove the screw [2] securing nozzle #6 to the mounting plate. Lift the nozzle straight up. 2. 3. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Replacement Action Steps Install Nozzle, #6, Blank 1. 2. 3. Reference Place nozzle #6 on the mounting plate and secure with one (1) screw [2]. Install the incoming and waste tubing [1] on the nozzle. Install the cuvette wash head cover. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Prepare for Operation 1. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.06 Nozzle, #7, Suction Version - 204738-101_4978_3 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.06 Nozzle, #7, Suction Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Nozzle, #7, Suction 1. Remove the vacuum tubing [1] from nozzle #7. Remove the screw [2] securing nozzle #7 to the mounting plate. Lift the nozzle straight up 2. 3. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Replacement Action Steps Install Nozzle, #7, Suction 1. 2. 3. Reference Place nozzle #7 on the mounting plate and secure with the one (1) screw [2]. Install the vacuum tubing on nozzle #7. [1] Install the cuvette wash head cover. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Prepare for Operation 1. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 0 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.07 Nozzle, #8, Dryer Version - 204738-101_4979_2 List/Part Numbers List/Part Number Description 2-89271-03 Nozzle, dry Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.07 Nozzle, #8, Dryer Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Removal Action Steps Prerequisite 1. The processing module must For further information, refer to the ARCHITECT System Operations Manual, Section be in Stopped or Offline status. 1: Use or function, Subsection: System status. Remove Nozzle, #8, Dryer 1. Remove the vacuum tubing [1] from the dry nozzle #8. Remove the dryer tip. Note Refer to Removal and Replacement Procedure R6.08 Cuvette Drying Tip Remove the two (2) screws [2] securing the dry nozzle #8 to the mounting plate. Lift the nozzle straight up. 2. 3. 4. Reference (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Replacement Action Steps Install Nozzle, 1. #8, Dryer 2. 3. Reference Place the dry nozzle #8 on the mounting plate and secure with the two (2) screws [2]. Install the vacuum tubing [1] on the dry nozzle #8. Install the cuvette drying tip. Note Refer to Removal and Replacement Procedure R6.08 Cuvette Drying Tip. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Prepare for Operation 1. 2. 3. 4. To change the status of the For information on the startup procedure, refer to the ARCHITECT System Operations processing module and sample Manual, Section 5: Operating instructions, Subsection: System startup, pause, and handler from Stopped to Ready, shutdown. perform the Startup procedure. Perform P-621 Cuvette Washer Alignment. Select System Maintenance, As Needed and perform 6052 Wash Cuvettes. Install the cuvette wash head cover. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.08 Cuvette Drying Tip Version - 204738-101_4980_2 List/Part Numbers List/Part Number Description 09D51-01 Cuvette Drying Tip 09D51-02 Cuvette Drying Tip 7-93428-01 Cuvette Dry Tip 1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.08 Cuvette Drying Tip Time Required 15 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Prerequisite 1. The processing module must be in For further information, refer to the ARCHITECT System Operations Manual, Section Stopped or Offline status. 1: Use or function, Subsection: System status. Remove Cuvette Washer Assembly 1. Loosen the black thumbscrew [1] until the cuvette washer assembly [2] can be lifted from the mounting plate [3]. Remove dryer tip 1. Caution Use gloves during this step. Residual oil interferes with the proper drying function of the tip. Lift the cuvette washer assembly up and rotate it so that the white dryer tip [1] is easily accessed. Note The cuvette washer nozzles are all attached to a black nozzle mounting plate. There is no need to remove any of the screws securing the wash nozzles to the Reference mounting plate. Remove the dryer tip [1] by pulling it off of the metal nozzle [2]. Discard the old tip. 2. 3. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Replacement Action Steps Install dryer tip and cuvette washer 1. 2. 3. 4. 5. Reference Gently install the new dryer tip. The wide side of the dryer tip faces the nozzles. Note The dryer tip and the Reaction Cuvettes are both rectangular. Install the dryer tip so that it fits into the cuvette. Position the bottom of the dryer tip 61mm ±0.5 mm from the underside of the cuvette washer assembly. Position the cuvette washer assembly on the alignment pins and tighten the black knurled knob. Perform P-621 Cuvette Washer Alignment. Select System Maintenance, As Needed and perform 6052 Wash Cuvettes. (If the video does not display, or to view the video full size: Click Here) Note Video contains no audio sound. Video Note C8000 cuvette washer nozzle used for demonstration of Removal and Replacement. However, instructions are identical for the removal and replacement of the c4000 cuvette washer nozzle. Prepare for Operation 1. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks 1 P-621 Cuvette Washer Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2014 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R6.09 Valve, DI Water Cuvette Wash Version - 204738-101_4981_1 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. R6.09 Valve, DI Water Cuvette Wash Time Required 10 min Tools/Materials Standard Tool Kit Absorbent Towel WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove Peristaltic Pump Head and Front Right Deck Cover 1. Turn the peristaltic pump head [2] counterclockwise (CCW) about 1/16 of a turn. Pull upward on the pump head to remove the head. Remove the front right deck cover [1]. 2. 3. Disconnect Tubing from Valve Note 1. 2. 3. Use absorbent towel to protect from spills. Clamp the tubing [1]. Unlock the hose clamps [2]. Remove the tubing [1]. Note The removal of the remaining tubing will be done later while removing the valve assembly. Remove the Valve 1. 2. 3. 4. 5. Remove Fittings 1. Disconnect the valve electrical plug [1] from the connector CWV. Remove the jack for connector CWV from the mounting bracket. Remove the two (2) screws [2] from the valve mounting bracket and remove the valve assembly. Remove the remaining tubing from the valve. Remove the two (2) screws securing the valve to the valve mounting bracket. Remove the fittings [1] from the ports of the valve. Replacement Action Steps Install Fittings and Valve 1. 2. Connect Tubing 1. 2. 3. 4. Install Valve and Bracket 1. 2. 3. Install pipe sealant tape on the threads of the fittings and install the three (3) fittings [1] onto the valve. Install the valve to the valve mounting bracket using two (2) screws. Connect the reinforced tube coming from the manifold to the barbed fitting on the valve. Connect the tubing [1] originating from the reagent supply centers "Y" fitting to the barbed fitting on the top of the valve. Connect the tubing from the cuvette wash pump to the valve, by screwing the connector to the threaded fitting. Secure the two (2) tubes with hose bands. [2] Install the valve and the bracket assembly to the frame using two (2) screws [2]. Install the jack [1] for connector CWV onto the mounting bracket. Connect the plug to connector CWV. Reference Install Peristaltic Pump Head and Front Right Rear Deck Cover 1. 2. 3. 4. Prepare for Operation 1. 2. 3. 4. 5. Install the front right deck cover [1]. Align the notch with the peristaltic motor tang. Align the attachment locking tabs on the peristaltic pump head [2] with the recesses on the pump. Turn head clockwise (CW) about 1/16 of a turn until it snap locks in place. Power ON the processing module. For information on the startup procedure, refer to the ARCHITECT Note The system control center power System Operations Manual, Section 5: Operating instructions, must be ON prior to turning on the Subsection: System startup, pause, and shutdown. processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the System icon. Perform 6052 Wash Cuvettes. Check for leaks. Verification Procedures Order VP Description VP Detail / Note 2 6052 - Wash Cuvettes Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R7.01 ICT Assembly Version - 204738-101_4982_1 List/Part Numbers List/Part Number Description 7-205387-01 ICT ASSY, C4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. R7.01 ICT Assembly Time Required 25 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the on board solution tray. Remove the Front Center deck cover [B]. Remove the Front Right deck cover [C]. Remove the Center Right deck cover [E]. 2. 3. 4. 5. 6. Remove ICT Module and Water Jacket. 1. 2. 3. 4. 5. 6. 7. Remove the ICT head top plate. Loosen the retaining screw and lift the ICT module [1] from the head. Disconnect the tubing. Disconnect the pre-amp cable from the ICT module. Remove the Allen set screw securing the ICT water jacket in the ICT housing [2]. Remove the two (2) screws securing the cover mounting plate to the module head mounting plate. [3] Remove the water jacket from the ICT Assembly. Remove ICT Assembly 1. 2. Disconnect the following cables from the ICT assembly: a. ISEAS from connector ISEAS b. IAAM from connector ISEAM c. IAEM from connector ISEEM. Remove the four (4) screws [1] securing the ICT assembly to the frame and lift the assembly straight up out of the processing module. Replacement Action Steps Install ICT Assembly 1. 2. Install ICT Water Jacket and ICT Module 1. 2. 3. 4. 5. 6. Prepare for Operation 1. 2. 3. 4. 5. 6. Reference Install the ICT assembly and secure it using with the four (4) screws. Connect the following cables: a. ISEAS to connector ISEAS b. IAAM to connector ISEAM c. IAEM to connector ISEEM Install the ICT water jacket with tubing into the ICT housing. Note Ensure the water jacket is against bottom of the ICT housing. Tighten the Allen set screw to secure the ICT water jacket. Install the cover mounting plate to the module head mounting plate using two (2) screws. Connect the tubing and the pre-amp cable to the ICT module and place into the ICT module head. Tighten the retaining screw. Install the head top plate. Install the Front Center deck cover [B]. Install the Front Right deck cover [C]. Install the Center Right deck cover [E]. Install the on board solution tray. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 3 6070 - Daily Maintenance 2 2134 - Change Water Bath 1 P-618 ICT Alignment N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Check for leaks Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. R7.02 Cable, ICT Pre Amp to ICT Version - 204738-101_4983_1 List/Part Numbers List/Part Number Description 2-89069-03 Cable, ICT pre amp to ICT Inspect tools for damage, ensure calibration is not expired and replace if necessary. R7.02 Cable, ICT Pre Amp to ICT Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard WARNING Probe Stick Hazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the center right cover [E]. Remove the front right cover [C]. 2. 3. Access Pre-Amp Cable 1. 2. Remove the ICT head top plate. Remove the retaining screw and lift the ICT module from the head. Remove Pre-Amp Cable 1. 2. Disconnect the ICT Pre-Amp cable. Follow the cable to the ICT Pre-Amp board and cut the cable ties as needed. Unscrew the pre-amp cable [1] from the pre-amp module. Remove the cable from the processing module. 3. 4. Replacement Action Steps Install Pre-amp Cable 1. 2. 3. Install ICT Module 1. 2. 3. Prepare for Operation 1. 2. 3. 4. Reference Screw the pre-amp cable to the connector on the pre-amp module. Route the cable to the ICT module and connect to the module. Secure the cable with wire ties. Place the ICT module into the ICT module head. Tighten the retaining screw. Install the head top plate. Install the center right deck cover [E] and secure using two (2) screws. Install the front right deck cover [C] using three (3) screws. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 2 G6 - Calibration Verify calibration per operator's manual. 1 6070 - Daily Maintenance N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. W1.01 Water Management Unit Version - 204738-101_4984_1 List/Part Numbers List/Part Number Description 7-205753-01 WMU, COMPLETE Inspect tools for damage, ensure calibration is not expired and replace if necessary. W1.01 Water Management Unit Time Required 90 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Turn off the incoming water source. Select the Fluidics/Wash category from the Diagnostics screen under the System icon. Perform 1175 Wash Cup Test. Note This relieves the water pressure in the incoming line. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover Remove the left cover. 2. 3. 4. 5. 6. 7. Remove Water Management Unit Cover and tubing 1. Drain Water Management Unit 1. 2. 2. Disconnect all tubing from the Water Management Unit (WMU). Remove the WMU cover. Drain the WMU using the service tubing [1]. Note The processing module drain tubing can be used to drain the WMU by placing the service drain inside of the drain tubing. If a container is used it must be able to hold a minimum of three (3) liters of water. When drained, replace the plug in the service drain. Disconnect Tubing 1. from DegasserWash Solution Pump 2. Manifold and Water Bath Drain Valve Disconnect the tubing to the Degasser at the Degasser-Wash Solution Pump Manifold [1]. Disconnect the tubing from the bottom fitting of the water bath drain valve. Remove Tubing from Low Concentration Waste Tank 1. Remove all tubing from the Low Concentration Waste (LCW) Tank as follows: a. From the vacuum pumps. [1] b. From the wash solution valves labeled 2 and 3. [2] c. Waste manifold. [3] Disconnect Cable Connections 1. Disconnect all cables from the connector panel on the left side. Remove Water Management Unit 1. 2. 3. 4. 5. 6. Remove the two (2) screws securing the WMU to the processing module frame. Disconnect the wash cup manifold tubing [3]. Note Tubing removed from one manifold must be installed on that same manifold. Mark as needed. Continue sliding the WMU to the rear. Disconnect the fitting [1] attached to the High Concentration Waste (HCW) ICT aspiration tubing. Disconnect the tubing from the HCW pump [2]. Slide the WMU out of the processing module. Replacement Action Steps Install Water Management Unit 1. 2. 3. 4. 5. Connect Cable Connections 1. Slide the WMU into the rear of the processing module far enough to allow the connection of the tubing. Connect the HCW tubing from HCW pump to the top of the middle waste tubing fitting at the rear of the water connection panel. Connect the fitting attached to HCW ICT aspiration tubing to the top waste tubing at the rear of the water connection panel. Connect the wash cup tubing to the respective wash cup manifolds (S, ST1 and ST2). Slide the WMU all the way into the processing module and secure using two (2) screws to the processing module frame. Note For more information see Water Management Unit Tubing Routing. Connect the following plugs to the connector panel: a. STFS b. RPFS c. SPFS d. STWV e. RPWV f. SPWV g. POWP h. WSCV i. FSDT j. FSUL k. FSLH l. FSLL Reference Install Tubing to Low 1. Concentration Waste Tank 2. 3. Connect Tubing to Degasser-Wash Solution Pump Manifold and Water Bath Drain Valve 1. Install Water Management Unit Cover and Tubing 1. 2. 2. 3. 4. Connect the tubing coming from the wash solution valves to the LCW. Connect the tubing from the vacuum pumps to the barbed fittings of the LCW. Insert the waste manifold tubing into the waste tank. Connect the tubing from the degasser to the Degasser manifold. Connect the tubing to the bottom fitting of the water bath drain valve. Install the WMU cover to the processing module. Connect the tubing as follows: a. HC Waste b. Overflow c. Drainage d. De-Ionized water inlet Install the left cover. Install the rear left cover, once the Wash cup test passes. Prepare for Operation 1. 2. 3. Turn on the incoming water source. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Note The processing module may require a second power on if the internal water tank does not completely fill during the first power on sequence. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks 3 1175 - Wash Cup Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. W1.02 Float Switch Assy Version - 204738-101_4985_1 List/Part Numbers List/Part Number Description 7-205547-01 WMU FLOAT SWITCH ASSY Inspect tools for damage, ensure calibration is not expired and replace if necessary. W1.02 Float Switch Assy Time Required 90 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Turn off the incoming water source. Select the Fluidics/Wash category from the Diagnostics screen under the System icon. Perform 1175 Wash Cup Test. Note This relieves the water pressure in the incoming line. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover. Remove the left cover. 2. 3. 4. 5. 6. 7. Remove Water Management Unit Cover 1. Drain Water Management Unit 1. 2. 2. Disconnect all tubing from the Water Management Unit (WMU). Remove the WMU cover. Drain the WMU using the service tubing [1]. Note The processing module drain tubing can be used to drain the WMU by placing the service drain inside of the drain tubing. If a container is used it must be able to hold a minimum of three (3) liters of water. When drained, replace the plug in the service drain. Disconnect Tubing 1. from DegasserWash Solution Pump 2. Manifold and Water Bath Drain Valve Disconnect the tubing to the Degasser at the Degasser-Wash Solution Pump Manifold [1]. Disconnect the tubing from the bottom fitting of the water bath drain valve. Remove Tubing from Low Concentration Waste Tank 1. Remove all tubing from the Low Concentration Waste (LCW) Tank as follows: a. From the vacuum pumps [1] b. From the wash solution valves labeled 2 and 3 [2] c. Waste manifold [3]. Disconnect Cable Connections 1. Disconnect all cables from the connector panel on the left side. Remove Water Management Unit 1. 2. 3. 4. 5. 6. Remove the two (2) screws securing the WMU to the processing module frame. Disconnect the wash cup manifold tubing [3]. Note Tubing removed from one manifold must be installed on that same manifold. Mark as needed. Continue sliding the WMU to the rear. Disconnect the fitting [1] attached to the High Concentration Waste (HCW) ICT aspiration tubing. Disconnect the tubing from the HCW pump [2]. Slide the WMU out of the processing module. Remove Float Switch 1. Assembly 2. From the connector panel, remove: a. FSLL (float switch level low) b. FSLH (float switch level high) c. FSUL (float switch upper limit) Note The connectors are not labeled. Mark the connectors when disconnecting from the connector panel. Remove the two (2) screws securing the float switch assembly to the water tank and lift from the water tank. Replacement Action Steps Install Float Switch Assembly 1. 2. Install Water Management Unit 1. 2. 3. 4. Install the float switch assembly on the water tank and secure it with two (2) screws. Insert the following connectors into the connector panel: a. FSLL (float switch level low) b. FSLH (float switch level high) c. FSUL (float switch upper limit) Slide the WMU into the rear of the processing module far enough to allow the connection of the tubing. Connect the HCW tubing from HCW pump to the top of the middle waste tubing fitting at the rear of the water connection panel. Connect the fitting attached to HCW ICT aspiration tubing to the top waste tubing at the rear of the water connection panel. Connect the wash cup tubing to the respective wash Reference 5. Connect Cable Connections cup manifolds (S, ST1 and ST2). Slide the WMU all the way into the processing module and secure using two (2) screws to the processing module frame. Note For more information see Water Management Unit Tubing Routing. 1. Connect the following plugs to the connector panel: a. STFS b. RPFS c. SPFS d. STWV e. RPWV f. SPWV g. POWP h. WSCV i. FSDT j. FSUL k. FSLH l. FSLL Install Tubing to Low 1. Concentration Waste Tank 2. Connect the tubing coming from the wash solution valves to the LCW. Connect the tubing from the vacuum pumps to the barbed fittings of the LCW. Insert the waste manifold tubing into the waste tank. 3. Connect Tubing to 1. Degasser DegasserWash Solution Pump 2. Manifold and Water Bath Drain Valve Connect the tubing from the degasser to the Degasser manifold. Connect the tubing to the bottom fitting of the water bath drain valve. Install Water Management Unit Cover Install the WMU cover to the processing module. Connect the tubing as follows: a. HC Waste b. Overflow c. Drainage d. De-Ionized water inlet Install the left cover. Install the rear left cover. 1. 2. 3. 4. Prepare for Operation 1. 2. 3. Turn on the incoming water source. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Note The processing module may require a second power on if the internal water tank does not completely fill during the first power on sequence. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. W1.03 Manifold, Wash Cup Valve Version - 204738-101_4986_1 List/Part Numbers List/Part Number Description 7-202646-01 Manifold, Wash Cup Inspect tools for damage, ensure calibration is not expired and replace if necessary. W1.03 Manifold, Wash Cup Valve Time Required 20 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Turn off the incoming water source. Select the Fluidics/Wash category from the Diagnostics screen under the System icon. Perform 1175 Wash Cup Test. Note This relieves the water pressure in the incoming line. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover. Remove the left cover. 2. 3. 4. 5. 6. 7. Remove Water Management Unit Cover 1. Drain Water Management Unit 1. 2. 2. Disconnect all tubing from the Water Management Unit (WMU). Remove the WMU cover. Drain the WMU using the service tubing [1]. Note The processing module drain tubing can be used to drain the WMU by placing the service drain inside of the drain tubing. If a container is used it must be able to hold a minimum of three (3) liters of water. When drained, replace the plug in the service drain. Remove Wash Cup Manifold 1. 2. 3. 4. Remove the bracket with the degasser manifold. Remove the tubing [1] on top of the Wash Cup manifold going to the valves. Remove the incoming water tubing. Remove the two (2) screws [2] securing the manifold to the frame and remove the manifold. Replacement Action Steps Install Wash Cup Manifold Valve 1. 2. Install the manifold and secure it with two (2) screws [2]. Insert the incoming water tubing. Reference 3. 4. 5. 6. 7. Prepare for Operation 1. 2. 3. Install the tubing [1] on top of the Wash Cup manifold going to the valves. Secure the bracket with the degasser manifold using the screws. Secure the service drain tubing to the WMU assembly frame using clamp. Install the WMU cover to the processing module. Install the rear left cover, once the Wash cup test passes. Turn on the incoming water source. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. Note The processing module may require a second power on if the internal water tank does not completely fill during the first power on sequence. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks 3 1175 - Wash Cup Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. W1.04 Flow Restrictor Version - 204738-101_4987_1 List/Part Numbers List/Part Number Description 7-93366-01 Flow restrictor, incoming Inspect tools for damage, ensure calibration is not expired and replace if necessary. W1.04 Flow Restrictor Time Required 35 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Turn off the incoming water source. Select the Fluidics/Wash category from the Diagnostics screen under the System icon. Perform 1175 Wash Cup Test. Note This relieves the water pressure in the incoming line. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover. Remove the left cover. 2. 3. 4. 5. 6. 7. Remove Water Management Unit Cover and tubing 1. Remove Flow Restrictor from WMU 1. 2. 2. Disconnect all tubing from the Water Management Unit (WMU). Remove the WMU cover. Remove the two (2) screws securing the Flow Restrictor bracket to the WMU frame. [1] Disconnect the tubing from the Flow Restrictor to the WMU. Note You may need to remove the tubing from the reservoir water tank. Remove Flow Restrictor from WMU (continued) 1. 2. Remove Flow 1. Restrictor, Incoming Water 2. 3. 4. Follow the cable from the solenoid and remove the WSCV connector from the connector bracket. Note Cut cable ties as needed. Remove the Flow Restrictor assembly from the analyzer. Remove the two (2) screws securing the valve to the bracket. Remove the on/off valve [1] and the flow restrictor with tubing attached. Remove the regulator tubing from the fitting [2] on the output of the flow restrictor. Unscrew the fitting [2] from the output of the flow restrictor. [3] 5. Unscrew the flow restrictor [3] from the fitting [4] between the on/off valve and the flow restrictor. Replacement Action Steps Install Flow Restrictor, Incoming Water 1. 2. 3. 4. 5. 6. Install Flow Restrictor on WMU 1. 2. 3. 4. Install Water Management Unit Cover and Tubing 1. 2. 3. 4. Cover the threads of the Flow Restrictor [3] with Teflon tape. Screw the flow restrictor [3] into the fitting [4] on the on/off valve [1] Cover the threads of the fitting [2] for the output of the flow restrictor. [3] Screw the fitting [2] into the flow restrictor. [3] Connect the tubing to the fitting [2] on the output of the flow restrictor and tighten the tubing clamp. Secure the on/off valve to the bracket using the two (2) screws. Secure the Flow Restrictor bracket to the WMU frame with the screws. Connect the tubing from the Flow Restrictor to the WMU and secure it with the clamp. Insert the cable from the solenoid onto the WSCV opening in the connector bracket. Connect the WSCV cable. Note Replace cable ties removed as necessary. Install the WMU cover to the processing module. Connect the tubing as follows: a. HC Waste b. Overflow c. Drainage d. De-Ionized water inlet Install the left cover. Install the rear left cover, once the Wash cup test passes. Reference Prepare for Operation 1. 2. 3. Turn on the incoming water source. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. Note The processing module may require a second power on if the internal water tank does not completely fill during the first power on sequence. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. W1.05 Valve, Incoming On/Off Version - 204738-101_4988_1 List/Part Numbers List/Part Number Description 7-93367-01 Valve, Incoming On/Off Inspect tools for damage, ensure calibration is not expired and replace if necessary. W1.05 Valve, Incoming On/Off Time Required 35 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Turn off the incoming water source. Select the Fluidics/Wash category from the Diagnostics screen under the System icon. Perform 1175 Wash Cup Test. Note This relieves the water pressure in the incoming line. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the rear left cover. Remove the left cover. 2. 3. 4. 5. 6. 7. Remove Water Management Unit Cover and tubing 1. Remove Flow Restrictor from WMU 1. 2. 2. Disconnect all tubing from the Water Management Unit (WMU). Remove the WMU cover. Remove the two (2) screws securing the Flow Restrictor bracket to the WMU frame. [1] Disconnect the tubing from the Flow Restrictor to the WMU. Note You may need to remove the tubing from the reservoir water tank. Remove Flow Restrictor from WMU (continued) 1. 2. Remove Incoming ON/OFF Valve 1. 2. 3. 4. Follow the cable from the solenoid and remove the WSCV connector [1] from the connector bracket. Note Cut cable ties as needed. Remove the Flow Restrictor assembly from the analyzer. Remove the two (2) screws securing the valve to the bracket. Remove the on/off valve [1] and the flow restrictor with tubing attached. Unscrew the fitting to the input of the on/off valve. Unscrew the fitting [2] from the output of the on/off valve to remove valve [1]. Replacement Action Steps Install Incoming ON/OFF Valve 1. 2. 3. 4. 5. 6. Install Flow Restrictor on WMU 1. 2. 3. 4. Install Water Management Unit Cover and Tubing 1. 2. 3. 4. Reference Cover the threads of the fitting for the incoming water. Screw the incoming water fitting into the on/off valve. Cover the threads of the fitting [2] with Teflon tape. Screw the fitting [2] into the on/off valve. [1] Connect the tubing to the fitting [3] on the output of the flow restrictor and tighten the tubing clamp. Secure the on/off valve to the bracket using the two (2) screws. Secure the Flow Restrictor bracket to the WMU frame with the screws. Connect the tubing from the Flow Restrictor to the WMU and secure it with the clamp. Insert the cable from the solenoid onto the WSCV opening in the connector bracket. Connect the WSCV cable. Note Replace cable ties removed as necessary. Install the WMU cover to the processing module. Connect the tubing as follows: a. HC Waste b. Overflow c. Drainage d. De-Ionized water inlet Install the left cover.. Install the rear left cover, once the Wash cup test passes. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Prepare for Operation 1. 2. 3. Turn on the incoming water source. For information on the startup procedure, refer to the ARCHITECT Power ON the processing module. System Operations Manual, Section 5: Operating instructions, Note The system control center power Subsection: System startup, pause, and shutdown. must be ON prior to turning on the processing module to ensure proper initialization. Note Note: The processing module may require a second power on if the internal water tank does not completely fill during the first power on sequence. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. W1.06 Pump, Mag. Wash Cups Version - 204738-101_4989_1 List/Part Numbers List/Part Number Description 7-205650-01 MAG PUMP, WMU, WASH CUP Inspect tools for damage, ensure calibration is not expired and replace if necessary. W1.06 Pump, Mag. Wash Cups Time Required 35 min Tools/Materials Standard Tool Kit Absorbent Towels WARNING Potential Biohazard Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Drain the water bath (System/Diagnostics/Modules/1101 Robotics Test Tool - Option 11). Turn off the incoming water source. Select the Fluidics/Wash category from the Diagnostics screen under the System icon. Perform 1175 Wash Cup Test. Note This relieves the water pressure in the incoming line. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the left cover. 2. 3. 4. 5. 6. Remove Water Management Unit Cover and tubing 1. 2. Drain Water 1. Remove the waste tubing, overflow tubing, and the incoming water line from the water connection panel. Remove the six (6) screws securing the WMU cover to the processing module and remove the cover. Drain the WMU using the service tubing [1]. Management Unit Note 2. Remove Magnetic Pump 1. 2. 3. 4. 5. The processing module drain tubing can be used to drain the WMU by placing the service drain inside of the drain tubing. If a container is used it must be able to hold a minimum of three (3) liters of water. When drained, replace the plug in the service drain. Remove POWP cable and connector from the connector bracket. [1] Remove the screw to release the ground cable from the Magnetic Pump. [2] Disconnect the tubing from the pump. [3] Remove the two (2) screws securing the pump to the WMU frame located on each side. [4] Remove the pump from the analyzer. Replacement Action Steps Install MagneticPump 1. 2. 3. 4. 5. Install Water Management Unit Cover and Tubing 1. 2. 3. 4. Prepare for Operation 1. 2. Reference Install the pump in position and secure to the WMU base with two (2) screws. Connect the tubing to the pump and secure it with the clamps. Insert the ground cable in the screw and secure it to the connector bracket. Mount the PWOP connector onto the bracket Connect the PWOP cable. Install the WMU cover to the processing module. Connect the tubing as follows: a. HC Waste b. Overflow c. Drainage d. De-Ionized water inlet Install the left cover. Install the rear left cover, once the Wash cup test passes. Turn on the incoming water source. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. Note The processing module may For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. 3. require a second power on if the internal water tank does not completely fill during the first power on sequence. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 2132 - Flush Water Lines Check for leaks 2 2134 - Change Water Bath Check for leaks 3 1175 - Wash Cup Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.01 RSH Carrier Transport Assembly Version - 204738-101_4990_3 List/Part Numbers List/Part Number Description 7-205729-01 CARRIER TRANSPORT ASSY, C4 AND CI41 Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.01 RSH Carrier Transport Assembly Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Open processing center cover. Remove the upper left side panel of processing module. Remove upper left panel support bracket. 2. 3. 4. VIDEO Remove Carrier Transport Assembly Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Carrier Transport Assembly 1. Move the carrier transport [1] to the far left position on the guide rail. 2. 3. 4. 5. 6. Remove transport rail stop on left end of guide rail Remove the carrier transport cover. Disconnect the flex cable connectors [2] at J2 and J3 on the carrier transport PCB. Remove the screw securing the retaining pin bracket [3] holding the flex cable to the carrier transport frame and remove cable. Remove the carrier transport from the RSH by sliding it off the guide rail. Replacement Action VIDEO Install Carrier Transport Assembly Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Carrier Transport Assembly 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. Slide the carrier transport onto the guide rail. Note Lift the transport assembly when middle roller is positioned in transport rail to prevent the base of the assembly from scraping the base of the RSH. Install the flex cable retaining pin bracket. Connect the flex cables connections to J2 and J3 on the carrier transport PCB. Install the transport rail stop. Install carrier transport cover. Install the upper left panel support bracket. Install upper left side panel. Power ON the processing module. Note The system control center power must be ON prior to turning ON the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G23 - Initialization/ Startup Verify instrument initializes and/or starts up. 2 1120 - Sample Pipettor Calibration 2 1114 - Carrier Transport Calibration 4 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.02 Carrier Transport Leadscrew with Lifter Assembly Version - 204738-101_4991_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.02 Carrier Transport Leadscrew with Lifter Assembly Time Required 30 min Tools/Materials #1 Phillips head screwdriver WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Shutdown 1. Processing Module Remove Carrier Transport Assembly 1. Remove Theta Shaft 1. The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). 2. Open processing center cover. 3. Remove the upper left side panel of processing module. 4. Remove upper left panel support bracket. 5. Remove transport rail stop on left end of guide rail. 6. Move the carrier transport to the far left position on the guide rail. 7. Remove the carrier transport cover. 8. Disconnect the flex cable connectors [2] at J2 and J3 on the carrier transport PCB. 9. Remove the screw securing the retaining pin bracket [3] holding the flex cable to the carrier transport frame and remove cable. 10. Remove the carrier transport from the RSH by sliding it off the guide rail. 2. 3. 4. Disconnect the cables at J1, J5, and J9 on carrier transport PCB. [6] Note Cut cable ties if needed. Remove two (2) screws [5] securing theta motor to transport frame. Carefully pull theta shaft arm from lead screw (L_screw) lifter. Slide theta shaft from transport frame. Reference For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Note Remove L_Screw from Carrier Transport Frame 1. 2. 3. Be careful not to damage the J9 cable when removing the theta shaft arm. Disconnect the Z motor cable at J8 and Z encoder cable at J6 [2] on the carrier transport PCB. Remove two (2) screws [3] securing the L_screw motor to frame. Remove the L_screw and motor. Replacement Action Steps Install L_Screw 1. 2. 3. Install Theta shaft Orient the L_screw with cables facing towards the left, and install the L_screw onto the carrier transport frame. Secure motor to frame with two (2) screws used from previous removal step. Connect the Z motor cable to J8 and the Z encoder cable to J6 on the carrier transport PCB. Replace removed cable ties. Note 1. 2. 3. 4. 5. Install Carrier Transport Assembly Reference 1. 2. 3. Use caution while handling to prevent damaging theta shaft. Do not handle by the shaft or turn by the flag at the top of the shaft. Position lifter towards center on the L_screw. Attach the theta shaft arm to the L_screw lifter before sliding theta shaft into top of carrier transport frame. Align pins on gearbox to holes on transport frame with motor power cables to the left. Secure theta shaft motor to carrier transport frame with the screws from removal steps. Connect cables J1, J5 and J9. Note Verify lifter moves up and down smoothly and arm turns freely. Slide the carrier transport onto the guide rail. Note Lift the transport assembly when middle roller is positioned in transport rail to prevent the base of the assembly from scraping the base of the RSH. Install the transport rail stop. Install the upper left panel support bracket. Prepare for Operation 4. 5. 6. 7. Install the flex cable retaining pin bracket. Connect the flex cables connections to J2 and J3 on the carrier transport PCB. Install carrier transport cover. Install upper left side panel. 1. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. 2. Verification Procedures Order VP Description VP Detail / Note 1 G23 - Initialization/ Startup Verify instrument initializes and/or starts up. 2 1114 - Carrier Transport Calibration 3 3240 - Bar Code Calibration 4 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.03 Carrier Transport Assembly PCB Version - 204738-101_4992_2 List/Part Numbers List/Part Number Description 7-205208-01 BD., CARRIER TRANSPORT, C4/CI4100 Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.03 Carrier Transport Assembly PCB Time Required 20 min Tools/Materials #2 Phillips Head Screwdriver WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Shutdown Processing Module 1. Remove Carrier Transport Assembly 1. Reference The processing module must be in Stopped For further information, refer to the ARCHITECT System Operations or Offline status. Manual, Section 1: Use or function, Subsection: System status. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). 2. Open processing center cover. 3. Remove the upper left side panel of processing module. 4. Remove upper left panel support bracket. 5. Remove transport rail stop on left end of guide rail. 6. Move the carrier transport to the far left position on the guide rail. 7. Remove the carrier transport cover. 8. Disconnect the flex cable connectors [2] at J2 and J3 on the carrier transport PCB. 9. Remove the screw securing the retaining pin bracket [3] holding the flex cable to the carrier transport frame and remove cable. 10. Remove the carrier transport from the RSH by sliding it off the guide rail. Remove Carrier Transport PCB Note 1. Turn theta shaft clockwise to verify the flag is not in a position to damage the sensor. Disconnect the following cables from the carrier transport PCB: • Carrier Transport Arm Sensor at J1 2. 3. 4. 5. • X-home sensor at J10 • Theta-motor encoder at J5 • Z-motor at J8 • X-motor encoder at J4 • Theta motor at J6 • X-motor at J7 • Ribbon cable at J9 • Carrier sensor J13 Remove the PCB cover. Remove the screws securing carrier transport PCB to transport frame. Remove two (2) posts for each of the connectors J2 and J3. Retain for use later. Remove PCB from transport frame. Replacement Action Steps Install Carrier Transport PCB 1. 2. 3. Install Carrier Transport Assembly 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. 2. Reference Install carrier transport PCB and secure to frame. Connect the following cables to the carrier transport PCB: • Carrier Transport Arm Sensor at J1 • X-home sensor at J10 • Theta-motor encoder at J5 • Z-motor at J8 • X-motor encoder at J4 • Theta motor at J6 • X-motor at J7 • Ribbon cable at J9 • Carrier sensor J13 Install two (2) posts for each of the connectors J2 and J3. Slide the carrier transport onto the guide rail. Note Lift the transport assembly when middle roller is positioned in transport rail to prevent the base of the assembly from scraping the base of the RSH. Install the transport rail stop. Install the upper left panel support bracket. Install the flex cable retaining pin bracket. Connect the flex cables connections to J2 and J3 on the carrier transport PCB. Install carrier transport cover. Install upper left side panel. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G23 - Initialization/ Startup Verify instrument initializes and/or starts up. 2 1114 - Carrier Transport Calibration 3 3240 - Bar Code Calibration 4 3310 - RSH Test N/A G110 - After repair is complete, verify per Expected Results released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.04 Carrier Transport Theta Shaft Version - 204738-101_4993_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.04 Carrier Transport Theta Shaft Time Required 30 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Shutdown Processing Module 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Carrier Transport Assembly 1. The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). 2. Open processing center cover. 3. Remove the upper left side panel of processing module. 4. Remove upper left panel support bracket. 5. Remove transport rail stop on left end of guide rail. 6. Move the carrier transport to the far left position on the guide rail. 7. Remove the carrier transport cover. 8. Disconnect the flex cable connectors [2] at J2 and J3 on the carrier transport PCB. 9. Remove the screw securing the retaining pin bracket [3] holding the flex cable to the carrier transport frame and remove cable. 10. Remove the carrier transport from the RSH by sliding it off the guide rail. Remove Theta Shaft and Motor Note 1. Turn theta shaft clockwise to verify the flag is not in a position to damage the sensor. Disconnect the following cables from the carrier transport PCB: • Carrier Transport Arm Sensor at J1 • Theta-motor encoder at J5 • Theta motor at J6 • Ribbon cable at J9 2. 3. 4. Remove two (2) screws [1] securing theta motor to transport frame. Carefully pull theta motor from support bracket. Release the theta shaft from Z mechanism lifter [2]. Replacement Action Steps Install Theta Shaft Reference Note 1. Use caution while handling to prevent damaging theta shaft. Do not handle by the shaft or turn by the disc at the top of the shaft. Insert the shaft into the transport frame. Note Make sure not to move the 2. 3. 4. Install Carrier Transport Assembly 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. 2. flag nor damage the sensor while placing the shaft in the top section of the RSH Carrier Transport. Position the shaft into the Z mechanism lifter. Note You can attach the theta shaft arm to the Z Mechanism lifter before sliding theta shaft into top of carrier transport frame. Secure theta shaft motor to carrier transport frame with the screws. Connect theta-motor encoder cable to PCB connector J5, theta motor at J6, ribbon cable to PCB connector J9 and carrier transport arm sensor cable to PCB connector J1. Note Verify lifter moves arm up and down smoothly and turns freely. Slide the carrier transport onto the guide rail. Note Lift the transport assembly when middle roller is positioned in transport rail to prevent the base of the assembly from scraping the base of the RSH. Install the transport rail stop. Install the upper left panel support bracket. Install the flex cable retaining pin bracket. Connect the flex cables connections to J2 and J3 on the carrier transport PCB. Install carrier transport cover. Install upper left side panel. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G23 - Initialization/ Startup Verify instrument initializes and/or starts up. 2 1114 - Carrier Transport Calibration 3 3240 - Bar Code Calibration 4 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.05 Carrier Transport Assembly X Motor Version - 204738-101_4994_2 List/Part Numbers List/Part Number Description 7-202962-01 Motor, RSH X-Axis, with Encoder Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.05 Carrier Transport Assembly X Motor Time Required 30 min Tools/Materials #2 Phillips Head Screwdriver WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Remove Carrier Transport Assembly 1. The processing module must be in Stopped or Offline status. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). 2. Open processing center cover. 3. Remove the upper left side panel of processing module. 4. Remove upper left panel support bracket. 5. Remove transport rail stop on left end of guide rail. 6. Move the carrier transport to the far left position on the guide rail. 7. Remove the carrier transport cover. 8. Disconnect the flex cable connectors [2] at J2 and J3 on the carrier transport PCB. 9. Remove the screw securing the retaining pin bracket [3] holding the flex cable to the carrier transport frame and remove cable. 10. Remove the carrier transport from the RSH by sliding it off the guide rail. Remove X-Motor from 1. Carrier Transport Frame 2. Disconnect the encoder and X motor cables from PCB connectors J4 and J7. Remove four (4) screws securing X-Motor to frame and remove x-motor. Replacement Action Steps Install X-Motor 1. 2. Install Carrier Transport Assembly 1. 2. 3. 4. 5. 6. 7. Prepare for Operation 1. 2. Reference Secure X-Axis motor to frame using the four (4) screws. Connect the encoder and X motor cables to PCB connectors J4 and J7. Slide the carrier transport onto the guide rail. Note Lift the transport assembly when middle roller is positioned in transport rail to prevent the base of the assembly from scraping the base of the RSH. Install the transport rail stop. Install the upper left panel support bracket. Install the flex cable retaining pin bracket. Connect the flex cables connections to J2 and J3 on the carrier transport PCB. Install carrier transport cover. Install upper left side panel. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G23 - Initialization/ Startup Verify instrument initializes and/or starts up. 2 1114 - Carrier Transport Calibration 3 3240 - Bar Code Calibration 4 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, Expected Results ENSURE appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.06 Carrier Transport Arm Sensor Version - 204738-101_4995_3 List/Part Numbers List/Part Number Description 7-97235-01 Sensor, Carrier Transport Arm 7-97235-02 Sensor, Carrier Transport Arm (RoHS) 7-97235-03 Sensor, Carrier Transport Arm (RoHS) Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.06 Carrier Transport Arm Sensor Note These videos represent simulated scenarios of instrument access and repair, and some may not accurately depict actual PPE guidelines. Trained personnel should follow procedures as outlined in Biological Hazards and safety procedures. Time Required 30 min Tools/Materials #2 Phillips Head Screwdriver WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Prerequisite The processing module must For further information, refer to the ARCHITECT System Operations Manual, Section be in Stopped or Offline 1: Use or function, Subsection: System status. status. Preparation 1. 2. 3. VIDEO Remove Carrier Transport Reference Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the upper left side panel of processing module. Remove upper left panel support bracket. Note Click the video and arrow icon in the reference section to access the video in Y2.01 RSH Carrier Transport Assembly action section: "VIDEO Remove Carrier Transport Assembly." Remove Carrier Transport 1. 2. 3. 4. 5. 6. VIDEO Remove Carrier Transport Arm Sensor Move the carrier transport [1] to the far left position on the guide rail. Remove transport rail stop on left end of guide rail. Remove the carrier transport cover. Disconnect the flex cable connectors [2] at J2 and J3 on the carrier transport PCB. Remove the screw securing the retaining pin bracket [3] holding the flex cable to the carrier transport frame and remove cable. Remove the carrier transport from the RSH by sliding it off the guide rail. Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Remove Carrier Transport Arm Sensor 1. 2. 3. 4. Disconnect the carrier transport arm sensor cable at J1 [1] on the carrier transport PCB. Remove the tie wrap securing the sensor cable to carrier transport frame. Remove the retaining pin securing the sensor cable to the carrier transport arm. [2] Pull the carrier transport arm sensor straight out from the carrier transport arm. CAUTION Do not pull hard on cable. It may be necessary to use a small blunt object (e.g. flat head screw drive) to push sensor out of arm. 5. Uncoil sensor cable [3] from theta shaft and remove. Replacement Action VIDEO Install Carrier Transport Arm Sensor Steps Reference Note Video contains no audio sound. (If the video does not display, or to view the video full size: Click Here) Install Carrier Transport Arm Sensor 1. 2. 3. 4. 5. Coil carrier transport arm sensor cable around theta shaft. Connect the cable to J1 on the carrier transport PCB. Insert the sensor into the carrier transport arm slot. Verify the sensor is pressed fully into the arm slot. Secure sensor cable to carrier transport arm using the retaining pin. Note 6. VIDEO Install Carrier Transport If retaining pin is damaged during the removal process, a tie wrap may be substituted. Tie wrap the sensor cable to the carrier transport frame. Note Click the video and arrow icon in the reference section to access the video in Y2.01 RSH Carrier Transport Assembly action section: "VIDEO Remove Carrier Transport Assembly." Install Carrier Transport 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. Slide the carrier transport onto the guide rail. Install the flex cable retaining pin bracket. Connect the flex cables connections to J2 and J3 on the carrier transport PCB. Install the transport rail stop. Install carrier transport cover. Install the upper left panel support bracket. Install upper left side panel Power ON the processing module. Note 4. The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 1 1114 - Carrier Transport Calibration 3 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009, 2015 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.07 Flex Cable with Pin Version - 204738-101_4996_2 List/Part Numbers List/Part Number Description 7-205489-01 FLEX CHAIN C4, INCLUDES CBLS, C4 7-205490-01 FLEX CHAIN CARRIER, C4 7-205491-01 FLEX CHAIN CBL, MTRS, C4 7-205492-01 FLEX CHAIN CBL, SNSRS, C4 7-205673-01 FLEX CHAIN, CI41, INCLUDES CARRIER AND CABLES 7-205675-01 FLEX CHAIN CBL, MTRS, CI41 Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.07 Flex Cable with Pin Time Required 30 min Tools/Materials #2 Phillips Head Screwdriver WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the screw securing the RSH distribution board cover and remove cover. Remove the four (4) screws securing the front cover assembly. Remove the two (2) screws securing the RSH front lower panel [1]. 2. 3. 4. Remove Flex Cable 1. 2. 3. Move the carrier transport to the far left position on the guide rail. Remove the carrier transport cover. Disconnect the flex cable connectors at J2 and J3 [2] on the carrier transport PCB. 4. 5. 6. 7. 8. Remove the screw [3] securing the retaining pin bracket holding the flex cable to the carrier transport frame. Disconnect the following cables from the rear of the RSH distribution board: • Flex cable motor signal at J13 • Flex cable data signal at J12 Remove the RSH Motor and Sensor cables from the cable clamp. Remove the two (2) screws securing the Flex cable bracket [4] to instrument frame. Remove flex cable from instrument. Note The flex cable is designed to flex only in one direction. If the cable is to be reinstalled, use care to ensure that the cable is not bent in the opposite direction. Replacement Action Steps Install Flex Cable Reference Note 1. 2. 3. 4. 5. 6. 7. 8. The flex cable is designed to flex only in one direction. If the cable is to be reinstalled, use care to ensure that the cable is not bent in the opposite direction. Connect the following cables to the rear of the RSH distribution board: • Flex cable motor signal at J13 • Flex cable data signal at J12 Secure the Flex Chain bracket to the instrument frame with the two (2) screws. Secure the RSH Motor and Sensor cables with the cable clamp. Route the flex cable into instrument and to the Carrier Transport Assembly. Install the RSH front lower panel. Install RSH cover. Secure the retaining pin bracket holding the flex cable to the carrier transport frame with screw. Connect the flex cable connectors at J2 and J3 on the carrier transport PCB. Prepare for Operation 9. Install the carrier transport cover. 1. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. 2. Verification Procedures Order VP Description 1 1114 - Carrier Transport Calibration 2 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.08 Carrier Transport X Home Sensor Version - 204738-101_4997_2 A Part Number is currently not assigned to this RR. Go to GPPM UserInterface Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.08 Carrier Transport X Home Sensor Time Required 15 min Tools/Materials #2 Phillips Head Screwdriver WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Shutdown Processing Module 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module. Remove X Home Sensor 1. Move the carrier transport to the far left position on the guide rail. Remove the carrier transport cover. Disconnect X-home sensor cable at J10 on carrier transport PCB. Loosen the screw securing the X home sensor mounting bracket to the carrier transport frame. Slide X-home sensor from transport. 2. 3. 4. 5. Replacement Action Steps Install X Home sensor 1. 3. Install X-home sensor mounting bracket onto transport carrier frame with cable oriented towards frame and secure with screw. Connect the sensor cable to J10 on the carrier transport PCB. Install carrier transport cover. 1. Power ON the processing 2. Prepare for Reference Operation 2. module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description VP Detail / Note 1 G23 - Initialization/ Startup Verify instrument initializes and/or starts up. 2 1114 - Carrier Transport Calibration N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y2.09 RSH Distribution Board Version - 204738-101_4999_2 List/Part Numbers List/Part Number Description 7-205203-01 BD., RSH DISTRIBUTION, C4/CI4100 Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y2.09 RSH Distribution Board Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the screw securing the RSH distribution board cover and remove cover. Remove the four (4) screws securing the front cover assembly. Remove the two (2) screws securing the RSH front lower panel [1]. 2. 3. 4. Remove RSH Distribution Board 1. 2. 3. 4. Disconnect the following cables from the RSH distribution PCB: • Barcode Reader at J5 • Load_Unload board cable W504 at J8 • RSH Motor Cable at J10 • RSH Sensor cable W0116 at J11 Loosen the four (4) green captive screws securing board to support mount frame. Remove RSH Motor and RSH Sensor cables from cable clamp underneath platform. Disconnect the two (2) cables from the rear of the RSH PCB: • Flex Cable Motor Signal at J13 5. • Flex Cable Data Signal at J12 Remove the RSH distribution board. Replacement Action Steps Install RSH Distribution Board 1. 2. 3. Install RSH Distribution board and secure board to mounting support frame with the four (4) green captive screws. Connect the following cables to the rear of the RSH distribution board: • Flex Cable Motor Signal at J13 • Flex Cable Data Signal at J12 Secure RSH Motor and RSH Sensor cables using the cable clamp. Reference 4. 5. 6. 7. Prepare for Operation 1. 2. Connect to following cables to the RSH distribution board: • Barcode Reader at J5 • Load_Unload board cable W504 at J8 • RSH Motor Cable at J10 • RSH Sensor cable W0116 at J11 Install the RSH distribution board cover. Install the RSH front lower panel. Install RSH front cover assembly. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 1114 - Carrier Transport Calibration 2 3240 - Bar Code Calibration 3 3310 - RSH Test N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y3.01 Bar Code Reader, RSH Version - 204738-101_4998_2 List/Part Numbers List/Part Number Description 7-205677-01 BCR, RSH 7-207194-01 Barcode Reader - Aggressive Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y3.01 Bar Code Reader, RSH Time Required 20 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Caution Class 2 Laser Radiation Removal Action Steps Reference Prerequisite 1. The processing module must be in Stopped or Offline status. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Preparation 1. Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Open the processing center cover to access the bar code reader. Note For an integrated analyzer, the barcode is located on the i1000 side 2. Remove Bar Code 1. Reader 2. 3. 4. Remove the one (1) screw that holds the board cover. Disconnect the cable [1]. Remove the two (2) screws [2] securing the bracket to the Bar Code Reader. Remove the two (2) screws securing the Bar Code Reader to the instrument frame. Note Retain screws for installation of barcode reader. Remove Bar Code 1. Reader (continued) 2. Release the cables from the cable clamps [1] and disconnect the Bar Code Reader cable [2]. Remove the two (2) screws securing the bracket to the Bar Code Reader. Replacement Action Steps Install Bar Code Reader (ci4100) 1. 2. 3. 4. Fix the Bar code reader to the instrument frame with the screws. Attach the Bar Code Reader to the bracket secure with two (2) screws. Connect it to the Bar Code Reader cable [2] and secure it the clamp [1]. Close the processing center cover. Reference Install Bar Code Reader (c4000) 1. 2. 3. 4. Prepare for Operation 1. 2. Attach the Bar Code Reader to the bracket secure with two (2) screws. Connect the Bar Code Reader cable. Install board cover and secure to the board with one (1) screw. Close the processing center cover. Power ON the processing module. Note The system control center power must be ON prior to turning ON the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. Verification Procedures Order VP Description 1 6070 - Daily Maintenance 2 3240 - Bar Code Calibration N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y3.02 Load/Unload Boards c4000 Version - 204738-101_5000_2 List/Part Numbers List/Part Number Description 7-205205-01 BD., LOAD/UNLOAD, C4/CI4100, C-SIDE 7-205206-01 BD., LOAD/UNLOAD, C4/CI4100, I-SIDE Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y3.02 Load/Unload Boards c4000 Time Required 20 min Tools/Materials Standard Tool Kit Caution Possible Electrostatic Discharge Shock Caution Class 2 Laser Radiation Removal Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. Preparation 1. 2. Remove Load/Unload Board 1. 2. The processing module must be in Stopped or Offline status. Shut Down the processing module. Remove the front lower cover to gain access to the board cover Remove the six screws that secure the board cover to the RSH. Separate the board to be replaced from the RSH Note While separating one board all may come down, since they are interconnected to each other at the end. Disconnect the ribbon cable from the left and the right of the board accordingly. Note The last board at the right has a terminator. Replacement Action Steps Replace Load/Unload Board 1. 2. 3. Connect the ribbon cables at both ends of the board. Mount the board in position. Note Assure all boards sit firmly in place. Install the board cover and secure it with Reference 4. Prepare for Operation 1. 2. the screws [1]. Note While tightening the screws press the board cover gently to ensure all boards stay in position. Replace the front lower cover and secure it with the screws. Note Ensure the cover sits properly with the hooks on the RSH. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 1 G1 - Verify the Repair according to released Operation and Service Procedures. N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. Y4.01 RSH Carrier Positioner Motor Version - 204738-101_5001_2 List/Part Numbers List/Part Number Description 7-205507-01 CAR POSN MTR_ENCODER Inspect tools for damage, ensure calibration is not expired and replace if necessary. Y4.01 RSH Carrier Positioner Motor Time Required 40 min Tools/Materials Standard Tool Kit WARNING Potential Biohazard Caution Possible Electrostatic Discharge Shock Removal Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. Preparation 1. 2. 3. 4. The processing module must be in Stopped or Offline status. Shut Down the processing module. Remove the Front center cover. [B] Remove the right cover. Remove the top screw [1] from rail stop; loosen the screw [2] from the bottom. Remove the carrier positioner from the rail. Preparation (continued) 1. 2. Remove the two (2) screws [3] that secure the positioner motor front cover. Remove the five (5) screws that secure the positioner rail [4]. Preparation (continued) 1. Remove the four (4) screws [5] that secure the Sample Wash Solution Holder. Remove Positioner Motor 1. Disconnect the cables [1] from the positioner board. Remove Positioner Motor (continued) 1. Remove four (4) screws [2] that hold the positioner motor to the bracket. Gently move the motor out of the mounting bracket. Note While removing the motor, be careful not to stretch the cables from the motor. When the motor is out, disconnect the cable [3] and cut the cable ties [4] as needed. 2. 3. Replacement Action Steps Replace Positioner Motor 1. 2. 3. 4. Connect the cable to the motor and secure the cables with cable ties. Insert the motor into the mounting bracket and secure using four (4) screws. Note While installing the motor ensure the cables are not crimped Connect the motor cables to the positioner board. Install the Sample Wash Solution Holder using four (4) screws. Reference 5. Install the positioner rail using five (5) screws. Note When placing the positioner rail, do not tighten the screws, ensure screws are flushed with the rail. 6. Insert the carrier positioner in the rail. Note While inserting the carrier positioner in the rail ensure the Teflon slider feet slides in the rail guide [1]. 7. Once the rack positioner is inserted, move it in front of the motor to ensure it moves smoothly, and then tighten the rail screws. 8. Return the rail stop to its original position, tighten the lower screw and insert and tighten the upper screw. 9. Replace the right cover. 10. Replace the positioner motor front cover and secure using two (2) screws. 11. Replace the front center cover. [B] Prepare for Operation 1. 2. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Verification Procedures Order VP Description 2 1120 - Sample Pipettor Calibration 1 1114 - Carrier Transport Calibration N/A G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. VP Detail / Note Expected Results ARCHITECT c4000 Service and Support Manual (Version 204733-103) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P1.02 Internal Covers Version - 201975-103_4349_2 Part Number Description 7-203133-01 7-97117-01 Cover, Processing Center (Internal) Cover, Processing Center Distribution Bd Inspect tools for damage, ensure calibration is not expired and replace if necessary. Time Required 5 min Tools/Materials Standard Tool Kit Removal Action Prerequisite Steps Reference 1. The processing module status must be in For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. Ready, Stopped, Offline or powered off. Remove Processing Center Cover 1. Raise the RV loader lower chute. 2. Press the latch tab [1] and raise the front of the cover, at the front edge of the processing center cover. 3. Lift the cover out of place. Caution Be very careful to avoid contacting the pipettor probe. Remove Distribution Board Cover 1. Loosen the three (3) captive screws at the top edge of the cover and lift the cover off. Replacement Action Install Distribution Board Cover Steps 1. Locate the two (2) bracket features. [2] Reference Install Distribution Board Cover (continued) 1. Install the distribution board cover, ensuring that the bottom edge fits inside the two bracket features. [3] Note The cover must NOT rest outside the bracket features. 2. Ensure that the top edge of the cover is properly seated and that the silk screening lines up with the indicator LEDs. 3. Tighten the three (3) captive screws on the top of the cover. Install Processing Center Cover 1. Position the rear edge of the processing center cover into place, carefully aligning the tabs at the back edge with the holes in the distribution board cover. 2. Lower the front of the cover and allow the latch to clip into place. [1] Note The cover should easily rest in position without interference and the front clip should latch easily. Verification Option Description Order VP Description 1 N/A VP Detail VP Note P-248 Startup and Shutdown G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. i1000SR Service and Support Manual (Version 201970-103) • © 2007, 2008 • ARCHITECT, i2000, i2000SR, c8000, ci8200, and c16000 are registered trademarks of Abbott Laboratories. i1000SR, i4000, i6000, i8000, and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-24 Precision Run Version - 201985-104_652_4 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-24 Precision Run Purpose The precision verification is used to verify that the instrument is functioning properly prior to releasing the system to the customer. Module i2000, i2000SR The procedure consists of calibrating and running specific assays with multiple replicates to test the pipetting, wash system, and optical precision of the instrument. The acceptance criterion is listed in the assay package insert and will determine if the instrument has passed or failed the verification. Materials (1 set or reagents) Required TSH B-hCG (1 set or reagents) Trigger/Pre-Trigger (1 Bottle each) Wash Buffer (The wash buffer bottle should be full) TSH Calibrators (1 box of Calibrator) B-hCG Calibrators (1 box of Calibrator) TSH Controls (1 Bottle of low Control) B-hCG Controls (1 Bottle of low Control) Assay Disk (1) Sample Carriers ( Set of at least 4 sequentially numbered carriers) RV/Sample Cups ( 1 bag each) Action Load Bulk Solutions and Reagents Verify TSB Status Steps 1. Verify that all bulk solutions are loaded and have been updated in the supplies screen. 2. Load the appropriate reagents. (TSH, B-hCG) 1. Verify that all required Technical Service Bulletins have been completed. Install 6114 Install/Delete Assays 1. Logon as FSE. 2. Select System Diagnostics. 3. Select the radio button (5) for module 5 (SCC). 4. Select Utilities tab, then select 6114 Install/Delete Assays. 5. Select F5 Perform, then OK followed by Install. 6. Select i2000, Proceed, All Groups. Note Enter the appropriate number for the following assays: TSH, BhCG (the numbers should be separated by a comma). Example: 3,11 7. Select Continue, Return to Previous Menu, Return to Main Menu, Exit, and Done. Configure Single Time Reference 90 min Analyte Control Precision File 1. At the snapshot screen select SYSTEM, CONFIGURATION. Select QC-Cal settings, QC-Single Analyte. 2. Select TSH and then select Configure F6. Note Use the arrow button to scroll down if needed. 3. Click on the Drop Down Menu Box next to Lot no: 4. Select NEW LOT and type Service Precision. Note Enter the information contained in table below in the New Level Screen. 5. Select Add Level then select Done. 6. Deselect the previously entered assay. 7. Repeat steps 2-5 for B-hCG. Perform Run Initialization 1. Select the SSH and the i2000 processing module on the Snapshot on screen. Processing 2. Select F8-Run. Module and Sample Note Handler To save time, while the processing module is performing the run initialization, enter calibration and control orders. Order Calibration Note Gather two consecutively numbered carriers. Consecutively numbered carriers are required for six point calibration assays. 1. Select ORDERS, CALIBRATION ORDER 2. Type in the Carrier Bar Code ID Number (C) and an unused Position Number (P). 3. Select Assay TSH, and B-hCG, then select F2-ADD ORDER. Note Use the arrow button to scroll down if needed. 4. Select Exit. Print the Order List and Load Calibrators into the Sample Carrier 1. Print the Order List to determine the minimum sample volume required per calibrator. 2. Select Orders from the Snapshot screen. 3. Select Order Status and then select Print F4. 4. Select Order List Report and then select Done. 5. Pipette the appropriate calibrator into the sample cups and load in the correct C/P(carrier/position) per the Order List Report. Note The Order List provides the correct amount of calibrator to be dispensed into each sample cup. 6. Place the Carrier onto the sample loader. Verify Calibration Order Precision Run 1. When the run has completed, select QC-Cal from the menu bar, and then select Calibration status. 2. Verify that the Cal Status for TSH, BhCG are listed as Active. 1. Select ORDERS, CONTROL ORDER then select SINGLE ANALYTE. 2. Type in the Carrier Bar Code ID Number (C) and an unused Position Number (P) 3. Select Assay(TSH), Lot(Service Precision), and Level(FSRPrec 1). Note Use the arrow button to scroll down if needed. 4. Select F5 ASSAY OPTIONS. Note In the number of replicates, UNDILUTED box enter 10. 5. Select Done and then select F2 ADD ORDER. 6. Repeat steps 3-6 for TSH. This will create a total of 20 samples for each assay. 7. Repeat steps 2-6 for twice for B-hGC. This will create a total of 20 samples for each assay. Note Carrier and position will be automatically assigned when each order is added. 8. Once all orders have been placed, print the Order List. 9. Select Orders from the Snapshot screen. 10. Select Order Status and then select Print F4. 11. Select Order List Report and then select Done. Load Low Controls into Carriers 1. Pipette the correct amount of low control for each assay and place the cups in the correct carrier positions, as per the printed Order List. Note The Order List provides the correct amount of control to be dispensed into each sample cup. 2. Place the Carriers into the sample handler. 3. If the modules are not in the Run mode select the SSH and i2000 and then select F8-Run. Note The Precision Control was created earlier during the Configure Single Analyte Control Activity. Print QC Summary Report Confirm Precision Data Results 1. Select QC Summary from the Snapshot screen. 2. Select the appropriate module number radio button. 3. Select (highlight) the assays that were run for Service Precision (levelFSRprec 1).Select Print F4. 4. Verify that the Selected Items radio button is selected. Select QC Summary Report, then select Done. 1. Confirm the precision results are acceptable using the Total %CV value listed in the appropriate assay package insert. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-201 Service Log On/Log Off Procedure Version - 96756-111_341_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-201 Service Log On/Log Off Procedure Purpose The purpose of this procedure is to document how to log on and log off as FSE. Log off is documented so that the system can be returned to normal operation after service or support. Module SCC Materials Required FSE Date Code List Time 1 min Log On as FSE Action Select Log On Log On Steps Reference 1. Select Log on F2 from the Snapshot screen. Note 1. Enter FSE in all capital letters in the User name field. 2. Enter the FSE logon date code in the password field. Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Note Enter the date code in reverse order. 3. Select Done. 4. Verify that the Operator listed on the Snapshot screen lists FSE. Log Off as FSE Action Select Log On Steps Reference Note 1. Select Log on F2 from the Snapshot screen. Note The User Name field is selected and populated with the name FSE. Prior to leaving the site, FSE is required to return the system to operator level of access. Log On as Operator 1. Press Delete to delete the FSE name. Note The password field disappears. 2. Select Done. 3. Verify that nothing is listed on the startup screen after Operator:. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-202 Modem Setup for Support Version - 96756-111_340_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-202 Modem Setup for Support Purpose Setup your system as the remote system for using PCANYWHERE to modem into ARCHITECT systems. Module SCC/Modem Materials Required PCANYWHERE Version 10.0 or higher Time Action Install PCANYWHERE Start PCANYWHERE Add Remote Icon Steps 1. Install per manufacturers instructions. 1. Select Start. 2. Select Programs. 3. Select Symantec PCANYWHERE. 1. Select the Remotes button. [1] 2. From the list of icons displayed, Double-click the Add Remote icon. [2] The PCANYWHERE Remote Properties: New Remote window displays. Select Modem 1. Verify the displayed modem is correct. If it is not correct, check the correct modem. 2. Select OK. The new remote icon displays with the name field highlighted. Reference 10 min Naming ARCHITECT Icon 1. Type in the name of your Remote Item (name of the account or generic as ARCHITECT). [1] 2. Press Enter. The new remote icon displays. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-203 SCC Diagnostic Software (Virtual PC-Check) Version - 96756-111_339_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-203 SCC Diagnostic Software (Virtual PC-Check) Purpose Instructions for using the SCC Diagnostic Software disk on the ARCHITECT SCC. Module SCC Materials Required Diagnostic Software, SCC (99058) Time Variable The SCC Diagnostic Software (Virtual PC-Check) is on any ARCHITECT SCC platform. Refer to the appropriate product code R&R and verification procedures or ISAs for the replacement and verification of SCC components. DO NOT REMOVE the PC-Check disk while running the program, unless requested by the program to insert a data disk. In that case, replace the program diskette immediately afterward. DO NOT ABORT during hard drive testing. This causes loss of system data. Normally, the screen display provides all information necessary to determine the nature of any fault detected. Action Steps Prerequisite 1. Exit the ARCHITECT application and Windows operating system. Note The PC-Check program is self-bootable and runs independently of the PC's operating system. 2. Verify that the SCC has a working system core, a functioning central processor unit, a floppy disk drive, and enough memory to load the program. 3. Complete processing or delete all pending test orders 4. Release all patient results 5. Archive all patient results 6. Release all QC results 7. Resolve all Exceptions and process Reruns, if necessary. Create New Backup 1. Create a new system backup. Note Use in the event of a catastrophic failure. Some tests may be destructive if the SCC power is interrupted during the testing process. 2. Perform diagnostic procedure 6004 (option 2) to copy the system backup to a CD. Start PCCheck Reference 1. Insert the PC-Check disk. 2. Perform a system shutdown and restart. If the system boots to the PC-Check disk, go to Select Desired Diagnostic. If the system does not boot to the PC-Check disk, access the BIOS setup function and verify that the floppy drive (removable device) is set up as the first boot device. Then, reboot with the PC-Check disk in the drive. Select Desired Diagnostic Note Refer to PC-Check Main Menu before completing this Action. 1. At the PC-Check main menu, select the desired diagnostic test. Note For most issues, the option Perform Basic Testing is sufficient to determine the cause of a problem. Note Passed/Failed results are displayed during this test option. Sufficient information should be given to determine which component is failing. Use this information to make component replacement decisions. 2. Select Exit (Write Reports) to exit PC-Check. 3. Replace the component(s) that failed according to the appropriate removal and replacement procedure 4. Perform the test option again for the indicated failure and verify a Passed result. If the problem is resolved by component replacement, go to Verify Replaced Component. Verify Replaced Component 1. Perform the appropriate verification procedure(s) as instructed in the removal and replacement procedure or ISA for the specific replaced component(s). PC-Check Main Menu Menu Item Description System Displays overview of system information, including motherboard, BIOS, processor, memory, cache, drives, video, Configuration audio, network, buses and ports. IDE Bus Information Scans IDE bus, accesses the devices found and checks and displays the configuration. Perform Performs a set group of tests (see PC-Check Test Groups) on the processor, motherboard, memory, cache, all Basic Testing drives, video adapter, printer functionality, speaker, keyboard and mouse. Interaction is required to test the mouse, keyboard and speaker. Passed/failed results are displayed. Note that several tests are not a part of the customized Abbott menu and will be displayed as `Not Run'. Other tests indicate `Absent' if specific hardware is not in the SCC. Note DO NOT attempt to run Basic Testing unless the appropriate amount of time can be allocated to complete all tests. ABORTING during hard drive testing will cause data loss and require that the base configuration and ARCHITECT application be reloaded on the hard drive. Approximate completion time is 20 minutes (completion time may vary depending on the platform type). For Platform D, two tests, Motherboard Realtime Clock Chip and CMOS RAM/Clock tests always show as FAILED due to an inherent incompatibility with the software. Disregard the FAILED results for these two tests when using the diagnostic software on a Platform D SCC. Note The default SCC keyboard type is Windows 104 key. The user determines appropriate selection. The printer port is LPT1. Selecting Generic for the printer type works in most cases. Perform Advanced Testing Performs memory inversion test, a hard drive read-write test and hard drive mechanics stress test. Approximate completion time is 20 minutes (completion time may vary depending on the platform type). Display Test Results Displays test results. Print Test Results Allows for printing of test results printed. About Virtual Lists PC-Check version and which components of PC-Check are active on this customized disk. PC-CHECK Exit (Write Reports Exits PC-Check. Does not write/save Results Summary. PC-Check Test Groups Group Test Basic Advanced Processor Core X Processor Coprocessor Core Processor X Processor Known Design Faults X Processor MMX Extensions X Motherboard DMA Controller X Motherboard System Timer X Motherboard Interrupt Controller Motherboard X Motherboard Keyboard Controller X Motherboard PCI Bus X Motherboard Real-Time Clock Chip X Memory Inversion Tree X Memory Progressive Inversion Memory X Memory Chaotic Addressing X Memory Block Rotation X Cache Memory Inversion Tree X Cache Memory Progressive Inversion Cache Memory X Cache Memory Chaotic Addressing X Cache Memory Block Rotation X Hard Drive Read X Hard Drive Non-Destructive Write Hard Drive X Hard Drive Mechanics Stress X Hard Drive Internal Cache X Video Adapter Text Memory Video Adapter X Video Adapter Super VGA Memory X Printer Printer Test X Speaker Speaker Test X Keyboard Keyboard X Mouse Mouse X i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-204 Log Retrieval Procedure Version - 96756-111_338_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-204 Log Retrieval Procedure This procedure documents retrieval of current logs, copying the database to CD, copying archived logs Purpose to CD, and retrieval of a Memory Dump file. If requested by Abbott personnel, the logs may be cleared from the system. Module SCC Retrieve current logs: This option copies the following logs to the D drive and floppy or CD if selected: Assay Activity, Assay Calibration, Assay Parameter, Assay QC Result Data, Liquid Level Sense, Maintenance History, Permanent Message History, Pressure Monitoring, Quality Control, Result, Scripting, Temporary, Temporary Message History, Sample Activity, Dr. Watson, Host, LAS, ARM, Arch, Unhandled, Application Event, CDRecord, Monitorlog, Printed Reports. In addition, the iSystem copies: Power On Self Test, PM Trace, Wash Aspirate Monitoring. In addition, the cSystem copies: Optics In addition, the i1000SR copies: Regeant Copy database to the CD: This option compresses and copies the database to the D drive and CD. Copy archive logs to CD: This option copies up to 5 previously retrieved compressed logs from the D drive to a CD. Retrieve memory dump file: (FSE only) This option compresses and copies a memory dump file to a CD. Refer to the tables below for additional information regarding each option. Materials Required Materials vary depending on the option selected. Required materials may include CD, or Floppies Action Steps Time Variable Reference P-201 Service Log On/Log Off Procedure Prerequisites 1. The system should be powered ON. 2. The processing module must be in the Stopped, Warming, or Ready state. Access 6009 Log Utilities Procedure 1. Select the System icon, then select Diagnostics from the drop- down menu. 2. Select the radio button for module 5. 3. Select the Utilities category. Retrieve Logs Note 1. Select 6009 Log Utilities procedure. Logs should not be cleared unless 2. Select Perform F5. A confirmation popup displays. 3. Select OK to perform the procedure. The diagnostic perform specifically requested by in-house Abbott personnel. dialog window displays. A description of the procedure displays on the instruction window. 4. Select Proceed to continue the log retrieval procedure. 5. Select the appropriate option. Refer to the tables below for additional information. 6. Follow the instructions on the screen. 7. When the procedure is complete, select Done. File Transfer (Optional) 1. Perform P-221 Modem File Transfer Procedure to transfer the logs via modem. Error Messages Error Message Probable Cause Corrective Action Not enough free space on A temporary copy of the files is copied to the C: drive before copying them to C: drive to create the CD. There is not enough space on the C: drive to copy the temporary files. temporary image files One of the following could have occurred: Non-ARCHITECT software was installed on the SCC. The C: hard drive contains bad sectors. Perform the appropriate drive recovery procedure Replace the hard drive No CD recorder detected The CD drive is not a read/write drive. The Gearworks drivers are corrupted. CD drive hardware failure Unknown error Perform the appropriate drive recovery procedure Replace the CD drive The COPY2CD_2 program was executed but did not respond. Perform the appropriate drive recovery procedure The COPY2CD_2 program has been moved or is corrupted. Retrieve Current Logs Option This option captures logs for troubleshooting or software investigation. The table below details each log and describes when that log is used. The log name has the following format: date_24time_serial#_logname.xxx (where x-axes is the extension of the log file). The files may be copied to a floppy disk, CDR/CDRW disk or hard drive. If floppy disk is selected the logs are compressed and saved to 1 or more floppies. If CD is selected the files are recorded uncompressed to a blank or appendable CDR or CDRW disk. If hard drive is selected the files are saved uncompressed to D:\Logs. In addition, all logs are saved in an archive file (refer to Copy Archive Logs to the C Drive option for additional information). Log Application Log Name eventapp.txt Steps Records the Log Format Format-specific for the application event Capacity Capture Log Module When Specific 9900 or 9077 No Event application event error information. error errors occurred and the SCC has shut down. Dr. Watson errors Records architect exception error information. Format-specific for architecture errors Architect exception errors ARM Issues Arch arch ARM debugcom7.log Records all debugcom7.log.bak communication between the SCC and ARM. Format-specific for ARM communication event. Assay Activity assayactivity Records for assay performance activities associated with performing tests Date/Time of Event | Module ID | Lockstep 24000 per Lockstep or Count | Lockstep Time | Test ID | module Timing errors Mechanism ID | Duration | Event | Error for i System or Code c System Assay Calibration assaycal Records for the last 600 attempted calibrations. Assay ID | Assay Status | Reagent Master 600 Lot | Calibrator Lot | Calibration Type | Completion Date | Completion Time | Module ID | Module SN | Operator ID | Curve Exception String | Curve Type | Calibration Summary | Curve Parameters | RMSE | Adjuster Ratios | Linear XFRM Slope | Linear XFRM Intercept Assay Parameter assayparameter Records a copy Formatted as an assay parameter report of the assay in assay numerical order. parameters for each assay installed. A copy of When you No the assay need to review parameter or verify the report for assay every parameter for assay an i System installed. and c System. Assay QC Result Log ReleasedQC.csv CVS format log Control Name | SID | Control Level | that contains all Control Lot Number | Control Comment | stored QC Assay Name | Assay Number | Module | results currently Serial Number | Cal Lot | Cal Date and in the database. Time | Reagent Master Lot | Reagent Serial Number | Date Time Completed | Carrier | Position | Result | Units | Min Range| Max Range | Dilution | Flags | Read Values | Absorbance Cuvette | Mv Cuvette | Operator ID | System Serial Number Stored QC results currently in the database. >Control No Issues for the i System and c System CDRecord CDRecord.log Captures a copy N/A of the log associated with the CD recording software. N/A Problems reading or writing to a CD-ROM. Dr. Watson user.dmp drwtsn32.log Records Dr. Watson event information. Format-specific for Dr. Watson events. debugcom5.log Records all See RS-232 Manual for format Host No No Cal or Result Issues for i System and c System Dr. Watson errors Host Issues Yes No No No debugcom5.log.bak communication between the SCC and Host. LAS debugcom6.log Records Date/Time of Event | Communication debugcom6.log.bak transferred Direction | Event between system and LAS Liquid Level lls Sense No information. 1000 LAS Issues No Records for fluid Date/Time of Event | Module ID | Test ID | 10,000 aspirate events Assay ID | Pipetting Context | Pipetting per including Step Number | Well Name | Level Sense module aspirate and Category | Level Sense Event | Step wickoff. Number Maintenance maintenancehistory Records for Date/Time of Event | Module ID | History maintenance Completion Status | Frequency | Operator performed. One ID | Event log is maintained for all modules. Level Sense Issues for i System and c System Maintenance or Diagnostic Procedure Issues for i System and c System Yes No Monitorlog (Note: only valid on SCC platform C and higher) Monitorlog.txt Fan and Date, Time, Error temperature monitor information from the intel monitor software. N/A When you No need to review errors from the intel monitor software. Optics optics Records the optics reads for a c8000 module. 10,000 per module Need to view absorbance data for wavelengths other than primary and secondary for c System 12000 Hardware or Software issues for i System and c System Permanent Message History messagehistory Power On Self Test post Pressure Monitor pm Replicate ID | Assay ID | Dilution ID | Sample ID | Test Completion Date | Test Completion Time | Module ID | Cuvette | Concentration | 16 wavelengths with 33 reads Records for the Date/Time of Event | Module ID | Error last 12000 error Category | Message Text codes logged in the message history category. Last record for Date/Time of Event | Module ID | Event each i System Generated processing module, SCC or sample handler power on sequence. Captures processing module, SCC, or sample handler bootup execution or error messages. Bootup Issues for i System or sample handler Records for pressure monitoring results of fluid aspirate events. Pressure Monitoring Issues for i System and c System Date/Time of Event | Module ID | Test ID | 10,000 Assay Number | Protocol | Step Number | per Well Name | Pipetting volume | PM Status module | Front End Pressure | Aspirate Pressure | Back End Pressure | Aspirate Error Sum | Yes No Yes Yes Actual Theoretical Ratio | Pressure Error | Stuff Printed Reports *.rtf An RTF version Formatted as the specific report of each report type. Quality Control qc Event recorded any time a user performs the following: includes or excludes a QC point, edits configuration parameters, changes Westgard rules, recalibrates reagent lot, edits QC lot default. Date/Time of Event | Module ID | Operator 5000 ID | Assay Name | Control Name | Control Lot Number | Control Level Name | Detail | Event QC event details for i System and c System Reagent Activity reagent Records the reagent activities associated with loading and unloading reagent carriers Reagent Carrier Detected I Reagent 4000 Carrier Load Complete I Reagent Carrier Unload Complete I Carrier Removed I Reagent Carrier Load I Reagent Carrier Unload I Reagent Carrier Error I Create Reagent Pack I Delete Reagent Pack I Report Mixing Status I Reagent Pack Last Aspiration I Disposition Reagent Pack I Carousel Window I Request Carousel I Request Carousel Response I Release Carousel I Reagent Pack Unload Time I Carrier Inventory I Aliquot Created I Reagent Pack Available Reagent carrier load and unload issues for c1000 Result result Records for the last 11,200 attempted tests run on the system. Replicate ID | Assay ID | Assay Status | 11200 Dilution ID | Sample ID | Sample Type | Patient ID | Test Order Date | Test Order Time | Test Initiation Date | Test Initiation Time | Test Completion Date | Test Completion Time | Assay Calibration Date | Assay Calibration Time | Module ID | Module SN | Operator ID | Dark Subreads | Signal Subreads | Dark Count | Signal Count | Corrected Count | Std Bak | Ave Bak | Std For | Ave For | Shape | Exception String | Result | Reported Result | Reported Result Units | Result Flags | Reagent Master Lot | Result Interpretation | Dilution Protocol | Result Details | Result Comment Result, LLS, and PM Issues for i System and c System Records all significant sample management events. Format-specific for each significant event. 1200 Sample Handler Issues Records script Format is specific for the maintenance or Sample Activity sampleactivity Scripting Script One RTF To obtain a version of copy of the each printed report. report type printed. 10,000 Need exists to No No Yes No Yes commands and responses from a maintenance or diagnostic procedure. diagnostic procedure. review specific errors that occurred when running the maintenance or diagnostic procedure Temporary temporary_*.log Records the log data into a temporary log as a backup. Temporary Message History tempmessage Records for the Date/Time of Event | Module ID | Error last 200 Category | Message Text messages logged in the temporary category (ie. Bar code read errors, host communication errors) 200 Trace trace Hardware Date/Time of Event | Module ID | Ticks | interactions to Event and from hardware devices (ie. Temperature controller, OMS boards) 10,000 per module No No Unhandled unhandled Records software error information. Format-specific for software error Wash Aspirate Mechanism wam Records for wash aspiration results of wash aspirate events. Date/Current_Time | Module_ID | Zone_Number | St1test# | St1Max | St1Delta | St2test# | St2Max | St2Delta | St3test# | St3Max | St3Delta Bar code read errors or Host communication errors for i System and c System Hardware issues for i System or sample handler No Yes No Non-specific software errors 3200 Wash Aspirate Issues for i System Yes Copy Database to CD Option This option backs up the database to the C: drive, compresses the database and records it to a CD. Once copied, the compressed file is deleted from the C: drive. This option may be required when requested for software investigation. Copy Archive Logs to CD Option This option allows the user to copy previously retrieved logs from the hard drive to a CD. When the Retrieve Current Logs option is selected, copies of the logs are compressed into one file and stored in D:\LOGS\ARCHIVEDLOGS. The file is named with the following format: date_24time_serial#_[1-5}.zip. Up to five compressed archived logs are saved on the hard drive with 1 being the newest and 5 being the oldest. The user is able to copy one or more archive files to a CD at a time. Retrieve Memory Dump File Option (FSE only) This option compresses and records the memory dump file to a CD. The memory dump file is used to gather data (debug information) on the utilization of system memory for the SCC. i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-205 Calibrate Touchscreen Version - 96756-111_337_3 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-205 Calibrate Touchscreen Purpose To calibrate the touchscreen on the SCC. Module SCC Materials Required None Time Action 3 min Steps Reference Access Explorer 1. Select the System icon. 2. Select Task Manager from the drop-down menu. 3. Select New Task. 4. Type Explorer. 5. Select OK. 6. Minimize Task Manager when Explorer opens. Calibrate Touchscreen 1. Using Explorer, scroll down and select Control Panel. 2. Double-click ELO Touchscreen to open it. 3. Select Align. 4. Follow the onscreen instructions to calibrate the touchscreen. 5. Select OK to complete the calibration. Note For some touchscreen monitors, better touchscreen response is achieved by performing the touchscreen calibration twice. Verify Touchscreen Calibration 1. Move through at least four screens, using the touchscreen, to verify touchscreen operation. i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-206 Printer Configuration Version - 96756-111_305_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-206 Printer Configuration Purpose Materials Required To configure the printer during installation, or when a new printer is installed. None Action Log On as FSE Time Steps SCC/Printer 15 min Reference Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Access Explorer Module 1. Select the System icon. 2. Select Task Manager from the drop-down menu. 3. Select New Task. 4. Type Explorer. 5. Select OK. 6. Minimize Task Manager when Explorer opens. Rename Note Architect If this is a new printer and no Architect Printer to Printer is listed, go to Load Printer Driver. Printer Product Name 1. Using Explorer, select the Printers folder. Note The printers folder is located in the Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. C:\Control Panel directory. 2. Right-click Architect Printer (highlights printer name) and select Rename from the drop down menu (selects printer name for editing). Note If a menu doesn't display when you right-click the name, select File, Rename from the menu bar. 3. Enter the printer product name (example OKI B4350). 4. Press Enter to rename the printer. Load Printer Driver Note If this is a new printer and the printer driver is already listed in the printers directory, skip this Action. 1. Insert the printer driver CD into the CD- ROM drive. Note If the CD starts automatically, stop that action. Exit the opening screen from the CD and return to the desktop. 2. Access the Printers folder and double- click Add Printer. 3. Select Next (starts the Add Printer Wizard). 4. Select the Local printer radio button. 5. Select Next. 6. Select LPT1 (connection port). 7. Select Next. 8. Select Have Disk. 9. Select Browse and locate the desired printer driver file on the CD-ROM. Note Refer to the appropriate ISA for the appropriate printer driver file name. 10. Select the driver and select Open. 11. Select OK, then Next several times until the Completing the Add Printer Wizard pop-up displays. Note Do not change any of the settings on these screens. 12. Select Finish. Note The Wizard loads the new driver. 13. Remove the printer driver CD from the CD- ROM drive. Rename Printer to Architect Printer 1. Right-click the name of the newly added printer (highlights printer name) and select Rename from the drop down menu (selects printer name for editing). Note If a menu doesn't display when you right-click the name, select File, Rename from the menu bar. 2. Enter Architect Printer as the new name. 3. Press Enter to rename the printer. Set Paper Size 1. Right-click Architect Printer and scroll down to Printing Preferences. Note If a menu doesn't display when you right-click the name, select File, Properties from the menu bar. 2. Select the Paper/Quality tab. 3. Select the Advanced button. 4. Select the desired paper size. 5. Select OK. 6. Go to Set Architect Printer as Default. Set Architect Printer as Default 1. Right-click Architect Printer (highlights printer name) and select Set as Default Printer. Note If a menu doesn't display when you right-click the highlighted item, select File, Set as Default Printer from the menu bar. 2. Verify that Set as Default Printer is checked in the menu that displays. 3. Repeat this step to verify that when you right-click the Architect Printer, a checkmark displays in front of Set as Default Printer. 4. Go to ARCHITECT Print Screen Verification. ARCHITECT Print Screen Verification Note 1. From any ARCHITECT screen, press Alt+Print Screen and verify that the screen prints. If this print test fails, refer to current ARCHITECT ISAs for information on Properties and Document Defaults for the printer. ARCHITECT Report Print Test Log Off as FSE Note 1. From the Snapshot screen, select the Results icon. 2. From the dropdown menu, select Results Review. 3. Select Print F4. 4. Select Results List Report. 5. Select Done to print. If this print test fails, verify that the printer is selected as the default printer. If it continues to fail, refer to current ARCHITECT ISAs for information on Properties and Document Defaults for the printer. 1. Select Log on F2 from the Snapshot screen. Note The User Name field displays and is populated with the name FSE. 2. Press Delete to delete the FSE name. Note The password field disappears. 3. Select Done. 4. Verify that nothing is listed on the startup screen after Operator:. Note Prior to leaving the site, FSE is required to return the system to operator level of access. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-207 ARCHITECT Print Test Version - 96756-111_354_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-207 ARCHITECT Print Test Purpose Verify printer is configured correctly. Module SCC/Printer Materials Required None Action Verify ARCHITECT Print Screen Verify ARCHITECT Prints Report Time Steps 3 min Reference Note 1. From any ARCHITECT screen, If this print test fails, refer to current ARCHITECT ISAs for information on Properties and Document Defaults for the printer. select the Alt+Print Screen keys and verify that the screen prints. Note 1. From the Snapshot screen, select the Results icon. 2. From the dropdown menu, select Results Review. 3. Select Print F4. 4. Select Results List Report. 5. Select Done to print. 6. Verify that the report header prints. If this print test fails, verify that the printer is selected as the default printer. If it continues to fail, refer to current ARCHITECT ISAs for information on Properties and Document Defaults for the printer. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-209 Use of Debug Cables for Troubleshooting (SCC Platform B or C) Version - 96756-111_293_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-209 Use of Debug Cables for Troubleshooting (SCC Platform B or C) Purpose To provide instructions on how to use debug cables to troubleshoot system start up issues. This procedure is especially useful when a processing module or sample handler remains in an OFFLINE status after adequate time for system start up has been allowed. Module SCC, i2000 SR, i2000, SH Materials 7-92394-01 Cables, Debug Kit Required Time Action Steps Prerequisite 1. The processing module and sample handler must be in a Stopped status. Preparation 1. Connect one debug cable from the top connector on the front of the card cage to connector P8 on the SCC DIGI board cable. Note This cable is for the processing module. 2. Connect one debug cable from the bottom connector on the front of the card cage to connector P7 on the SCC DIGI board cable. Note This cable is for the sample handler. Prepare Software 1. Power down the processing module. 2. Log on as FSE (P-201 Service Log On/Log Off Procedure). 3. Select System and then select Task Manager. 4. Select New Task. 5. Enter Explorer and then select OK. 6. Navigate to C:\TERMINALS. 7. Double-click on SSH.HT. The sample handler hyperterminal window displays. 8. Double-click on IAHV.HT. The processing module hyperterminal window displays. 15 min Reference Observe System Status During Start Up 1. Power on the processing module. Note After power is applied, data displays in the processing module and sample handler hyperterminal windows. 2. Allow the start up process to proceed until an error occurs or until the processing module and sample handler are in a Stopped status. 3. Review the data generated during the start up process for errors or other problems. i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-210 Modem Test Version - 96756-111_292_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-210 Modem Test Purpose To be performed on installation of a system SCC, or modem. Module SCC/Modem Materials Required None Action Verify System is Ready Modem Into System Time 15 min Steps Reference 1. Verify system power is ON. 2. Verify modem is connected and ON. 3. Switch (Alt+Tab) to the PCANYWHERE window and verify status is Waiting. 1. Contact Country Support to dial in using PCANYWHERE Version 10.0 (or higher), or use a separate computer equipped with PCANYWHERE (the FSR laptop) and a modem connection to dial into the SCC. Verify Remote User Connection 1. Remote user verifies that they are able to connect to the SCC and see the ARCHITECT user interface. 2. Remote user verifies that they can move through at least four different screens without losing connection. 3. Close remote user connection. Verify System is Ready 1. Switch (Alt+Tab) to the PCANYWHERE window and verify the status has returned to Waiting. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-211 System Date, Time, and Time Zone Version - 96756-111_291_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-211 System Date, Time, and Time Zone Purpose Set the date, time, and time zone and enable or disable daylight savings time. Module SCC Materials Required None Action Log On as FSE Time Steps 5 min Reference Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Set Date, Time, and Time Zone 1. Select the System icon. 2. Select Configuration from the dropdown menu. 3. Select the System settings radio button. 4. Select the System control center category. 5. Select Configure F6. 6. Enter the System date and System time. 7. Select the desired time zone. Note Place a check in the Automatically adjust clock for daylight saving changes checkbox to select automatic adjustment for Daylight Savings Time. 8. Select Done. Verify Settings 1. Access the Snapshot screen. 2. Verify the date and time are correct. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-212 PCANYWHERE Startup Version - 96756-111_290_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-212 PCANYWHERE Startup Purpose This procedure is used to configure PCANYWHERE to manual startup (no modem installed), reconfigure automatic startup after a modem is installed, and restart PCANYWHERE. Module SCC/Modem Materials Required None Time Action Steps Log On screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. 1. Select the System icon. 2. Select Task Manager from the drop- down menu. 3. Select New Task. 4. Type Explorer. 5. Select OK. 6. Minimize Task Manager when Explorer opens. Access Control Panel 1. Using Explorer, scroll down and select Control Panel. 2. Select it to display the icons on the right side of the screen. Access Services Reference Note 1. Select Log on F2 from the Snapshot Access Explorer 15 min 1. Scroll down the Explorer window to find Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. the Administrative Tools icon. 2. Double-click the Administrative Tools icon. 3. Double-click the Services icon. 4. Verify that the Services window opens. 5. Scroll down to find PCANYWHERE Host Service and double-click the icon to select it. PCANYWHERE Settings 1. View the Service status and the Startup type. The Startup type [1] indicates PCANYWHERE status during power ON. Automatic indicates PCANYWHERE starts automatically at power ON. Manual indicates the user must start PCANYWHERE in the Services Control panel The Service status [2] indicates if the service is or is not active. Started in the status field indicates that the PCANYWHERE service is running. Stopped indicates that the PCANYWHERE service is not running. Disable PCANYWHERE Note Perform this procedure if a modem is not installed on the SCC. 1. Set to Manual Startup: From the services window, select PCANYWHERE Host Service from the list. Set the Startup type to Manual. [1] Select OK to save the setting. 2. Stop the service: If the Service status indicates Started, select Stop to stop the PCANYWHERE Host Service. [2] A message displays indicating that the system is attempting to stop the PCANYWHERE service. 3. Select OK to save the setting. 4. Close the Services window. 5. Restart the SCC and verify that PCANYWHERE does not start automatically. Verify/Configure to Automatic Startup and Start the Service This is only required if PCANYWHERE does not go to the waiting mode when the SCC is restarted. This may occur if the service setting for PCANYWHERE was changed from Automatic to Manual start. Note Verify that a modem is connected, turned on, and the telephone line is functional. Action Verify/Configure to Automatic Startup Steps Reference 1. In the Services window, with PCANYWHERE Host Service selected, double-click. 2. Select the Automatic Startup type. Start Service 1. If the Service status lists Started, verify the modem is operational according to P- 210 Modem Test. 2. If the Service status lists Stopped, select Start to start PCANYWHERE Host Service. A message displays attempting to startup PCANYWHERE and then puts it in Waiting. Note If an error message is received select OK. 3. Select OK to save the setting. 4. Close the Services window. 5. Restart the SCC. Verify PCANYWHERE Starts Automatically 1. Switch (Alt+Tab) to the PCANYWHERE window and verify the status is Waiting. Verification 1. Verify modem operation using P-210 Modem Test. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-213 Modem Connection to an SCC Version - 96756-111_289_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-213 Modem Connection to an SCC Purpose This procedure documents how to use PCANYWHERE to modem into an ARCHITECT SCC. Module SCC/Modem Materials Required None Time Action Contact Customer Steps 1. Contact the customer and obtain permission to modem into their system. 2. Inform them why you are modeming in and what activities you expect to perform while connected to their system. 3. Ask the customer to verify that the modem is on and connected to a telephone line. Verify PCANYWHERE Status 1. Have the customer press Alt+Tab to locate the PCANYWHERE icon. Note If the icon is not available, then PCANYWHERE is not open or is not in a waiting status. 2. Have the customer verify that the PCANYWHERE dialog indicates PCANYWHERE in a waiting status. If PCANYWHERE is waiting, go to Open PCANYWHERE. If PCANYWHERE is not waiting, ask the customer to restart the SCC, or refer to P-212 PCANYWHERE Startup. The customer may have closed PCANYWHERE accidentally PCANYWHERE has been changed from Autostart to Manual start. Open PCANYWHERE 1. Start PCANYWHERE. Reference N/A 2. Select the Remotes button. [1] 3. Double-click the ARCHITECT icon. [2] Note If you work with pre-configured accounts, double-click the account icon and the computer automatically dials the number. Enter Modem Number 1. Enter the modem number for the ARCHITECT SCC. [1] Be sure to enter any necessary access codes (for outside line, etc.) 2. Select OK. Log In 1. Enter FSE in the Username field. 2. Enter fse in the Password field. 3. Enter SCC in the Domain field. Note The PCANYWHERE file manager window displays. Once connected to an SCC, an image of the SCC screen displays on your system. You can perform: All on-screen operations. Most keyboard operations. Note Response to mouse clicks and keyboard entries is slower than when you are working directly on an SCC. Also, the colors used in the ARCHITECT software are not correctly presented in the PCANYWHERE screen image. This does not impact available functionality. Complete Modem Activities 1. Refer to the functions and activities listed under Accessing Different Applications on the SCC to complete the call. If file transfer is required, refer to P-221 Modem File Transfer Procedure. Close Call 1. If you logged onto the system as FSE at any time during your call, log off at this time (refer to P-201 Service Log On/Log Off Procedure.) 2. Select the End Remote Control Session icon. [1] 3. Contact the customer to let them know that you have completed your modem activities. 4. Document your activities according to local country call management procedures. Accessing Different Applications on the SCC Note The ALT+TAB key combination cannot be used to move between applications from a remote computer. Use Windows NT Task Manager to switch between programs, end programs, or launch new programs on the SCC. Action Log On as FSE Steps Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Launch Task Manager 1. Select the System icon on the SCC image. 2. Select Task Manager from the drop-down menu. Switch Programs Reference 1. Open Task manager. 2. Select the program you want to use. 3. Select the Switch To button. [1] Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Close Application 1. Open Task manager. 2. Select the program you want to end. 3. Select End Task. [1] 4. Select Wait in the next dialog. Note If the same dialog displays and the application did not quit, select End Task in the dialog. Open Programs 1. Open Task manager. 2. Select the New Task button. [1] If you know the name and the path of the application you want to start, you can type it in the dialog box. [2] If you want to select the application by browsing, select Browse [3] and find the program in the Browse window. 3. Select OK. 4. Minimize the Task manager window. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-214 Database Restore Version - 96756-111_288_3 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-214 Database Restore Purpose If the system has database errors or will not boot up, perform this procedure to restore the database from the system software. Module SCC Materials Required None Time 20 min Restore from Utilities Action Log On as FSE Steps Reference Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Restore Database 1. Select the System icon, then select Utilities. 2. Select the Backup Software radio button. 3. Select a backup from the available backup list. 4. Select [F5] Restore and restore the: Module Calibration System Configuration Database 5. The system restarts and the module calibration, system configuration, and database are restored. 6. For i1000SR, after restoring a system backup, force the system to unload all the reagents from the carousel and then reload them. Module Calibration - System calibration files, robo.cal and sample handler.cal files. System Configuration Default host and system parameters. Database - System database file, Architect.bak. Contains M&D procedures, assay files, assay calibrations, results, module and SCC configuration. This may be done one of two ways: 1. Through the user interface, from Snapshot screen, select Reagents > Select all > F7 Unload. Once the processing module is set to Ready and the RSH to Running, the reagents will automatically unload. Then, remove and reinsert the carriers to reload them. OR 2. While the processing module and RSH are in Stopped or Ready status, fully open and close the reagent carousel cover, which will cause all the reagents to go to BCR Fail status. Once the processing module is set to Ready and the RSH to Running, the reagents will automatically unload. Then, remove and reinsert the carriers to reload them. Restore at Boot Up (cannot start ARCHITECT Software) Action Access Restore Window Restore Database Steps Reference 1. When the system software starts to bootup and you get a screen that has the ARCHITECT logo, press Alt+Shift. 2. Verify that a restore window displays. 1. Select the desired backup file (the file to restore) to highlight it. 2. Select the Create database button. 3. Select Done. Verification 1. Verify the ARCHITECT application starts correctly. 2. Perform startup on the processing module and the sample handler. 3. Update Inventory. i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-215 PM Log Interpretation Version - 96756-111_287_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-215 PM Log Interpretation Purpose Use to interpret PM log data after performing procedure P-204 Log Retrieval Procedure. Module SCC Materials Required Floppy, CD, or modem for retrieval Time N/A Example of PM Log: The above sample represents only some of the potential errors. More than one error can occur, such as the 0X0150, because they are or'd (combined) together. The 0X00150 shows a communications error and clot (0X0010) and two shape errors (0X0020). PM Status Definition 0X0001 Invalid Syringe Probable Cause and Corrective Action PMIB is telling Syringe to cut on/off too early. Try PMIB, LLS Board, Software protocol, experiment protocol 0X0002 Air Only Aspirated No liquid was picked up during aspiration. Also known as front end pulse error. Check LLS to see if it is triggering above the liquid Check for droplets on side of container (condensation, splashing) Is the container empty? Check for connection tightness, and then start with sensor 0X0010 Clot Detected Clot Score Exceeded 1000. A clot was sucked in during aspiration. Bad Samples. Check sample, especially if whole blood, for chunky bits (clots). Replace if clumpy. Try a new probe then a sensor. Be sure to flush using CLI command for the pipettor (OMS XX JGXXXX) before running PM again. 0X0020 Shape Error Shape Error. Air was aspirated after liquid was sampled. Shape score in log exceeds 180 Sample fluid not homogenous (usually blood samples) Probe hitting side of container. Check Robo Cal Erratic LLS. Check LLS log for consistent liquid found position Can be loose robot components (mount screws or Z-Motion Slop) 0X0040 Start /Stop 0X0100 Communications Suspect when you get multiple different errors on the same pipettor. Can be combined with another Error number. 0X0500 Communications The system may not see all the data. Error Broken or open connections on PMIB board. LLS/PMIB Connector wires broken. Check connections or replace PMIB Replace LLS Board It has been observed that the PM/LLS components sometimes work fine except in certain combinations with a particular pipettor. Try swapping the whole set of parts between a working pipettor and the one in question. 0X0200 Over Pressure Detected Overpressure detected. For overpressure, either the probe or sensor is clogged or the probe hit the bottom of the container and sealed off without z-crashing. 0X0400 Under Pressure Under pressure detected. Detected For under pressure, check for liquid in container General If you get a large number of errors on a single pipettor regardless of sample size, the probable causes are: Sensor (erratic, not linear any more, leaking) PMIB Connections LLS Board i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-216 WAM Log Interpretation Version - 96756-111_286_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-216 WAM Log Interpretation Purpose Use to interpret WAM log data after performing procedure P-204 Log Retrieval Procedure. Module SCC Materials Required Floppy, CD, or modem for retrieval Time N/A A guideline to use for the WAM Station Delta at each probe position is that it should be at least 1250 units (1.25 C (33.6 F) degrees). The WAM Log record structure (each line of data) is: Date Current_Time Module_ID Zone_Number St1test# St1Max St1Delta St2test# St2Max St2Delta St3test# St3Max St3Delta 06/10/1998 17:36:43 |1 |2 | 25 | 27274 | 34 |0 | 27649 | 852 |0 28808 | 2651 St1, St2, St3 means station 1, 2, 3 in that order. Note In the logfile, it is all squeezed together to conserve disk space and looks like: 06/10/1998 17:36:43|1|2|25|27274|34|0|27649|852|0|28808|2651 Note Aspirations occur twice per lockstep - this means two lines of data for each lockstep. Note If no RV is present at a probe position, data is reported, but is very low. This occurs as an RV enters and moves through a wash zone. At the first lockstep when an RV enters a wash zone, the data for Probe 1 looks normal, while the data for probe 2 and 3 is low. At lockstep 2, the data for probe 1 and 2 should look normal, while the data for 3 is low. At lockstep 3, the data for all three probes should look normal. (This assumes that there are at least three (3) RVs in the process path.) i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-217 Disable Keyboard Locale Hot Keys Version - 96756-111_275_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-217 Disable Keyboard Locale Hot Keys Purpose This procedure documents how to disable the hot keys for local languages. Module SCC Note If locale hot keys are enabled when Alt+Shift keys are selected by the operator, the operating system switches from US (standard) keyboard mapping to other keyboards (such as German, French, Spanish, and Italian). Materials None Required Time Action Log On as FSE Steps Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Disable Keyboard Locale Hot Keys Reference Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Access Explorer 2 min 1. Select the System icon. 2. Select Task Manager from the drop-down menu. 3. Select New Task. 4. Type Explorer. 5. Select OK. 6. Minimize Task Manager when Explorer opens. 1. Use Explorer to navigate to the Control Panel folder. 2. Double-click on Control Panel. 3. Double-click on Keyboard. 4. Select the Input Locales tab. 5. Select Change Key Sequence button. 6. De-select Enable Key Sequence check box. 7. Select OK. 8. Select Apply. 9. Select OK. Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. 10. Close Explorer. Log Off as FSE Note 1. Select Log on F2 from the Snapshot screen. Note Prior to leaving the site, FSE is required to return the system to operator level of access. The User Name field displays and is populated with the name FSE. 2. Press Delete to delete the FSE name. Note The password field disappears. 3. Select Done. 4. Verify that nothing is listed on the startup screen after Operator:. Verification 1. Press Alt+Shift on the keyboard. Note It is important that you only press Alt+Shift once. If Alt+Shift is pressed more than once, the keyboard mapping is not correctly set for the verification. 2. Select the Orders icon. 3. Select Patient order from the drop-down menu. 4. Go to the SID field and press y on the keyboard. Note If "z" displays, repeat the procedure. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-218 C: Drive Recovery Procedure (SCC Platform C) Version - 96756-111_276_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-218 C: Drive Recovery Procedure (SCC Platform C) Purpose This procedure is used when the integrity of the system software (SCC Platform C) is suspect (incorrect shutdown of the SCC is a possible cause) or to install software when replacing the C: drive. This procedure includes making a backup of the database, rewriting the C: drive, reconfiguring the SCC, and restoring the database. Module SCC Materials ARCHITECT System Software Configuration Media Required Time Action Steps 30 min Reference Prerequisite 1. The processing module must be in Stopped, Ready, or Offline status. Log On as FSE Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Verify Current Backup Note 1. Select the System icon, then Utilities. 2. Select Backup software from the Utilities screen. 3. Verify that a current backup displays in the Available Backups field. Insert CD and Restart SCC 1. Insert the ARCHITECT System Software Configuration Media in the CD-ROM drive. 2. Select Shutdown [F3] from the Snapshot screen. 3. Select OK to confirm the Shutdown request. 4. Press Ctrl+Alt+Del keys simultaneously. 5. When the Confirm exit window displays, select Shutdown and restart the computer and then select OK. If this cannot be performed, go directly to Insert CD and Restart SCC. Note When the SCC restarts, a WARNING! You are about to install ... message displays. 6. Press any key to pass the WARNING screen. Note The R:\> prompt displays. Rewrite C: Drive Note 1. At the R:\> prompt, type 2 (Software Restore for C:) REWRITE platform... "C", Drive C: and press Enter. 2. Press Enter to pass the WARNING! screen. Note The system starts rewriting and the C: drive is rewritten with a clean installation of the operating system, application programs, and default system files. 3. When prompted to insert the next media, remove Disk 1 of 2, insert Disk 2 of 2, and select Enter (continues the process). 4. When the rewriting process is complete and the R:\> prompt displays, remove Disk 2 of 2 from the CD-ROM drive. 5. Reboot the computer by simultaneously pressing the Ctrl+Alt+Delete keys. Note The system restarts. When the System Settings Change window displays, the version splash screen displays 2.01 rather than 2.02 because the changes made between software version 2.01 and 2.02 affect only the c8000 module firmware. 6. When the "Do you want to restart your computer now?" displays, select Yes. Note This reboot ensures all of the hardware is recognized before continuing. Note If the display graphics mode is not correct, select Shutdown [F3] to shutdown and then restart the SCC to correct the condition. Configure First Module Note When the SCC reboots, the Configure first module window displays. 1. Press the Num Lock key to activate the number keypad. 2. Enter the FSE logon code. 3. Select the first processing module type from the list. Note The ARCHITECT Software Configuration CD is a service tool and should be taken from the site when the service engineer leaves. If installing a ci8200, the first processing module is c8000. 4. Enter the Serial No. of the first module being installed. 5. Select OK to shutdown and restart the SCC. Note After the SCC reboots a new database is created. This may take up to 20 minutes. When complete, the Snapshot screen displays. 6. When 0537 Improper shutdown on SCC error message displays, select OK. Note If a modem is not installed, the software displays a Dr. Watson error. Perform procedure P-212 PCANYWHERE Startup to set the service to manual. Verify Software Version 1. When the Snapshot screen displays, log on as FSE (P- 201 Service Log On/Log Off Procedure). 2. Select the System icon, then Utilities. 3. Select Software install and verify software version 2.02 displays. Restart SCC and Locate Backups 1. From the Utilities screen, select Backup software. 2. Verify that the last backups display in the Available Backups field. 3. Locate the backup that you want to restore. Restore Database 1. Select the database backup that you wish to restore. 2. Select Restore, and restore the: Module Calibration - System calibration files, robo.cal and ish.cal files. System Configuration - Default host and system parameters. Database - System database file, icw.odb. Contains M&D procedures, assay files, assay calibrations, results, module and SCC configuration. 3. The system restarts and the module calibration, system configuration, and database are restored. Log On as FSE Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Access Explorer 1. Select the System icon. 2. Select Task Manager from the drop-down menu. 3. Select New Task. 4. Type Explorer. 5. Select OK. 6. Minimize Task Manager when Explorer opens. Calibrate Perform P-205 Calibrate Touchscreen. Touchscreen Configure Printer Identify and Perform Required Upgrades Perform P-206 Printer Configuration. Note 1. Determine if the software version installed requires any upgrades. Compare the TSB sticker on the SCC and the processing module to the latest version of the TSB/ISA listing. Use Explorer to view the folders on the D: drive. If a Change or Upgrade folder exists, and has content, an upgrade has been performed. Reinstalling software on the C: drive erases any upgrades performed after the installation of the latest software version using the Software Configuration CD-ROM. It is necessary to reinstall those upgrades. 2. Perform any required upgrades per the instructions documented in the TSB or ISA. Startup Processing Module(s) and Update Inventory Perform Control Run 1. Perform startup on the processing module(s) and sample handler. 2. Update Inventory. 1. Using the reagents that shipped to the site, calibrate and run at least two (2) sets of 10 (total of 20) replicates of each level of control for one Two Step Assay Protocol. 2. Verify that each of the results are within the ranges published. 3. Print the QC report and verify that the CVs are under 10%. (Refer to P-24 Precision Run) Note Use all positive results for the calculation. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-219 Ethernet Communication Verification Version - 96756-111_294_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-219 Ethernet Communication Verification Purpose This procedure verifies that Ethernet communication has been established between the SCC and the Module SCC processing module(s) or sample handler. Materials Required None Time Action Steps 15 min Reference Power ON SCC 1. Power ON the system control center. Note Do not go to the next step until the system control center displays the Snapshot screen. Power ON the Module(s) 1. Power ON the processing module(s). 2. Verify that the status of the attached module(s) goes from Offline to Stopped. Note During data transfer, the corresponding LED on the Ethernet hub blinks rapidly. Perform Startup 1. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-220 LLS Log File Interpretation Version - 96756-111_253_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-220 LLS Log File Interpretation Purpose Use to interpret LLS log data after performing procedure P-204 Log Retrieval Procedure. Module SCC Materials Required Floppy, CD, or modem for retrieval LLS Log Field Definitions Log excerpt showing how LLS should have increasing steps going into a container: Regular LLS Log segment without errors: LLS Status Codes Time N/A Readouts generated by the LLS X Status command: Note Works better if LLS X CLR_STATUS is run first, where X is the LLS system number, 0 (sample), 1 (R1), or 2 (R2). Liquid Sensed, first time after replacing the cpu, or LLS board 0909 Liquid sensed, normal readout 0101 Lost contact during aspiration 0505 Self test readout, never saw liquid 0808 Shifted 0101 readout, may have crashed the probe 1011 Positively crashed the probe 1010 Theta Ground (only shows up during robo cal) 0606 i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-221 Modem File Transfer Procedure Version - 96756-111_281_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-221 Modem File Transfer Procedure Purpose Use to transfer files to and from an ARCHITECT SCC. This allows you to retrieve system logs, verify system files, or download updated files to the customer SCC. Module Modem Materials Required None Time Action Access SCC PCANYWHERE File Manager Steps N/A Reference 1. Select the File Transfer icon. [1] Locate File(s) 1. Locate the file(s) you want to transfer on the remote system or on the host system. Transfer File(s) 1. Open the folder to which you want the file(s) transferred. 2. Select the file(s) you want to transfer. 3. Select the Transfer icon. [1] Note The folder D:\Service\Temporary is on every SCC as a temporary folder for files that you want to transfer. The folder is a temporary storage location; once file transfer is complete and the files are used appropriately, delete the files in this folder. Reference: Verify Transfer 1. Verify that the file(s) that you transferred displays in the receiving folder. 2. Complete the call according to P-213 Modem Connection to an SCC. The remote system (your PC) is on the left-hand side of the screen. [2] The host system (SCC) is on the right hand side of the screen. [3] i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-223 Override Processing Module Cover Interlocks Version - 96756-111_280_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-223 Override Processing Module Cover Interlocks Purpose Module SCC Time NA This procedure is used to override the following covers: front processing module cover (c4000, i1000SR, i2000SR, i2000) rear processing module cover (i2000SR, i2000) R1, R2, and sample carousel (c8000, c16000) reagent supply center, reagent supply center maintenance (Y) (c4000) Materials Required None Action Log On as FSE Steps Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Override cover interlocks Reference 1. Select System from the Snapshot screen. 2. Select Configuration, and then Modules. 3. Select the desired processing module, and then select On for Override interlocks. Note Results are not reported when the cover interlocks are overridden. WARNING The automatic reagent carousel scan is not Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. performed if you open and close the reagent carousel covers. If any reagents or segments are removed or changed you must request a manual reagent scan. Change override cover interlocks to Off 1. Select System from the Snapshot screen. 2. Select Configuration, and then Modules. 3. Select the desired processing module, and then select Off for Override interlocks. Note Prior to leaving the site ensure that the override cover interlocks is configured to Off as results are not reported when the covers are overridden. i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-226 Set Bar Code Transitions Version - 96756-111_214_5 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-226 Set Bar Code Transitions Purpose Materials Required This procedure is used to reduce the frequency of the invalid bar code read errors by editing the transition field of the appropriate bar code SCC Module calibration file to a value correct for the bar code type, presentation, or number of characters being used. Example Sample Tubes with customer bar code labels, pretreatment tubes or other tubes with special labels. Action Steps Time Reference Prerequisites 1. Power ON the system. 2. The processing module must be in the Stopped, Warming, or Ready state. Preparation 1. Obtain a variety of sample bar codes to be used. Note Some sites use a variety of bar code types and number of characters. To ensure that the file is set properly to read all tubes, it is advised to determine the minimum number of characters used. 2. Find out the bar code type (symbology) that is used. Log On as FSE Determine Number of Characters Perform 3253 Bar Code Configuration 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. 1. At the Snapshot screen, select Orders and then Patient order. 2. Select the SID field. 3. Using the SCC bar code scanner, scan the bar code label on a tube. 4. Count the number of characters entered into the field. 5. Repeat the process for the next tube. 6. Of the labels scanned, determine the minimum number of characters. 1. Perform 3253 Bar Code Configuration, Transition Value option (System, Diagnostics, SCC Option, Utilities), and go to Shutdown and Restart SCC. Manually Set Note Number of The following steps are an optional method to manually enter transition values to the BCR.Cal Transitions File. This is required for values greater than those shown in the Bar Code Transitions table. Use (Optional) minimum value that gives reliable results. 1. Using Bar Code Transitions, locate the bar code symbology and the number of characters. 2. Determine the recommended number of transitions. Note Barcode types serve as examples. Special labels may use transition settings closest to those shown on the table for the example type. All settings are verified prior to use by performing the appropriate test listed at the end of this procedure. Both tube and empty positions are checked. Transitions as high as 120 would be acceptable. Bar Code Transitions 15 min Access Explorer Open Bar Code Cal File 1. Select the System icon. 2. Select Task Manager from the drop-down menu. 3. Select New Task. 4. Type Explorer. 5. Select OK. 6. Minimize Task Manager when Explorer opens. 1. Navigate in Explorer on the left side to the desired folder for the type of sample handler. RSH and/or LSH (Sample Carousel) = C:\IPLS\RSH\System iRSH (c4000, i1000SR, ci4100) = C:\IPLS\iRSH\System SSH = C:\IPLS\ISH\System LAS = C:\IPLS\LASB\System 2. In the right field, select and double-click on BCR.CAL. 3. If the Open With pop up screen displays, select NOTEPAD. Edit Transition Value 1. In the BCR.CAL file, locate the value for the transition value number under the Value column. 2. Change the number to the new value determined above. 3. Edit only the numerical value. DO NOT edit or delete any of the characters. Note The information for the Microscan MS3 barcode reader is not used at this time. This information will only be present on instruments at v5.00 software or higher. If an error occurs during the editing process go to Restore Default File. Save File 1. At the top of the screen, select FILE then SAVE. Restore Default File Note This action is only required if there was an error in editing the BCR.CAL file. 1. Navigate in Explorer on the left side to the desired folder for the type of sample handler. Note Example: For the RSH you would navigate to C:\IPLS\RSH. 2. In the right field, select and double-click on the default directory and select BCR.CAL. 3. From the menu, select EDIT then COPY. 4. Navigate in Explorer to the appropriate sample handler system directory. 5. From the menu, select EDIT then Paste. 6. Go to Open Bar Code Cal File. Shutdown and Restart SCC 1. Perform P-248 Startup and Shutdown to shutdown and restart the computer and processing modules. 2. Go to Perform Diagnostic Procedure Bar Code Reader Test. Perform Diagnostic Procedure Bar Code Reader Test 1. Perform the appropriate Maintenance and Diagnostic procedure SSH - 3250 SH Bar Code Reader Test RSH - 3252 RSH Bar Code Reader Test iRSH (c4000, i1000SR, ci4100) - 3230 Bar Code Reader Test LAS - 3255 LAS Crsl Bar Code Reader Test LSH / Sample Carousel - 3251 Sample Carousel Bar Code Test For information on the startup procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. Note Use at least 10 bar coded tubes from the site when performing Maintenance and Diagnostics. 2. Verify that all the bar coded tubes used scan correctly. i System Service and Support Manual (Version 96661-116) • © 2002, 2010 • ARCHITECT, i1000SR, i2000, i2000SR, ci4100, c8000, ci8200, c16000, ci16200 are registered trademarks of Abbott Laboratories in various jurisdictions. c4000, i4000, i6000, i8000 and ARCHITECT ARM are trademarks of Abbott Laboratories in various jurisdictions. • Abbott Park, IL 60064 • All rights reserved. P-230 D: Drive Recovery Procedure Version - 96756-111_328_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-230 D: Drive Recovery Procedure Purpose This procedure is used when the integrity of the D: drive is suspect (incorrect shutdown of the SCC is a possible cause) or to install software when replacing the D: drive. Module SCC Materials Required ARCHITECT System Software Configuration Media Time Action Insert CD and Restart SCC Steps 20 min Reference 1. Insert the ARCHITECT System Software Configuration Media in the CD-ROM drive. 2. Select Shutdown [F3] from the Snapshot screen. 3. Select OK to confirm the Shutdown request. 4. Press Ctrl+Alt+Del keys simultaneously. 5. When the Confirm exit window displays, select Shutdown and restart the computer and then select OK. Note When the SCC restarts, a WARNING! You are about to install ... message displays. 6. Press any key to pass the WARNING screen. Note The R:\> prompt displays. Rewrite D: Drive Note 1. At the R:\> prompt type 3 (Software Restore for D:) REWRITE platform... , Drive D: and press Enter. 2. Press Enter to pass the WARNING! screen. Note The system starts rewriting and the D: drive. 3. When the rewriting process is complete and the R:\> prompt displays, remove the CD from the CD-ROM drive. 4. Reboot the computer by simultaneously pressing the Ctrl+Alt+Delete keys. Note The system restarts. Wait until the System Settings Change window is displayed. The ARCHITECT Software Configuration CD is a service tool and should be taken from the site when the service engineer leaves. 5. Select Shutdown [F3] and then select OK to confirm the shutdown request. 6. When the "Do you want to restart your computer now?" displays, select Yes. Note This reboot ensures all of the hardware is recognized before continuing. Identify and Perform Any Required Upgrades Note Reinstalling software on the D: drive erases any upgrade information saved in the Change or Upgrade folder on the D Drive. The TSB sticker can be used to obtain information on required upgrades. 1. Determine if any upgrades affect the information just installed on the D Drive. Compare the TSB sticker on the SCC and the processing module to the latest version of the TSB/ISA listing. 2. Perform any required upgrades that affect the D: Drive per the instructions documented in the TSB or ISA. Backup Database 1. When the Snapshot screen displays, select the System icon, then Utilities. 2. Select Create backup [F4] (performs database backup). 3. Verify that the backup occurs without error. Startup the Processing Module(s) and Update Inventory 1. Perform startup on the processing module(s) and sample handler. 2. Update Inventory. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-231 Software Error Information Version - 96756-111_329_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-231 Software Error Information Purpose Checklist for gathering software error information. This information should be gathered and documented in Module SCC the call management system, when a software error occurs. This checklist should be completed for any software error, which reoccurs several times or requires onsite service. Materials None Required Description Time Information Required Describe the error in detail (list all actions taken, all system responses). If possible, print a screen image of the error as not all errors are logged in the message history and temporary message. List all related error messages (message history and temporary message). How often does this error occur? Error Description Did the error occur while printing a report? If so, which one? Did the error occur while printing a screen image (Alt + Print Screen)? Did the error occur while accessing the SCC? Did the error occur while creating patient test orders, calibrators test orders, or controls test orders? What keystrokes were being selected when the lockup or error occurred? Check for Hidden Error Messages A hidden error message may be present. Hold down the Alt+Tab keys and look for the following Icon. Use the Tab key to tab to this icon and deselect the Alt key. If an error message is present, it should display. What was the status of the Processing Module at the time of error? Status and Peripheral Devices What was the status of the Sample Handler at the time of error? Can the operator use the touchscreen? Keyboard? Mouse? Does the SCC respond to STOP on the touchscreen or keypad? 20 min List Information Here What is the module configuration (i2000, i4000, i2000SR, c8000, etc.)? What is the Software Version and when was it installed? (System, Utilities, Software Install) Were there any power issues at the time of the lockup? Does the customer have a UPS? Is it correctly plugged into the backup outlet? (P-42 UPS Installation) If the system is configured for a Host interface, what was the system doing at the time? Was it querying the host or downloading tests from the host? System Configuration & Activities If the system is configured for a Host interface, was it releasing patient, QC, or exception results to the host? If the system is configured for a Host interface, was it scanning patients or controls? Was it running bar coded controls? If the system is configured for automatic wash buffer transfer using the ARCHITECT ARM, was an automatic or a manual transfer in progress? If the system is configured for automatic wash buffer transfer using the ARCHITECT ARM, was a Host transfer occurring at the same time as the transfer or ARCHITECT ARM request? Log Collection Perform Diagnostic Procedure 6009, Log Utilities (System, Diagnostics, Module 5, Utilities, 6009 Log Utilities, Option 1, Retrieve Current Logs, Select Log Destination - Floppy, CD, or Hard drive.) Note Up to five (5) sets of logs can be stored to the Hard drive. The logs can be copied to a CD at a later time using 6009, Option 3, Copy Archived Logs to CD. The logs can also be sent via modem at a later time. Database Collection (if needed) Collect the Database for the following issues: Reagent Issues (mismatches, read errors, etc.) Assay Configuration Issues (such as Retest rules) Search Function Issues Tests not transitioning properly in Order Status Screen or Results Review Screen (tests staying in Scheduled, Pending Transmission) Copy the Database to a CD using 6009, Log Utilities, Option 2, Copy Database to CD Rebooted the SCC? Recovery - What procedure was used to recover from the error? Rebooted the Processing Module? Restored a backup database? Reinstalled software? (P-218 C: Drive Recovery Procedure (SCC Platform C)) i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-232 Delete Reagent Kit Version - 96756-111_330_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-232 Delete Reagent Kit Purpose This procedure documents deletion of a reagent kit from the system. Module SCC Materials Required None Action Time Steps Log On as FSE 5 min Reference Note 1. Select Log on F2 from the Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Access Reagent History Screen 1. Select the Reagent icon from the Snapshot screen. 2. Select Reagent history from the drop down menu. Delete Reagent Kit Note 1. Select the desired reagent kit. 2. Select Delete - F6. 3. Repeat Step1 and Step2 to delete additional kits. 4. Verify deletion of the selected reagent kit(s). To delete an assay instead of a reagent kit, perform diagnostic procedure 6114 Install/Delete Assays. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-234 System Serialization Version - 96756-111_333_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-234 System Serialization Purpose Materials Required To configure serial numbers in the system software at installation. None Time Action Locate Serial Number Configuration Information Module Steps 1. Architect System No. - Completed by the SCC vendor. 2. SCC Serial Number - Completed by the SCC vendor. Note The first character of the SCC Serial Number is an alpha character. It identifies the hardware used in the SCC so that the correct replacement parts and Hard Drive Images can be installed. 3. SH Serial Number: RSH (c8000, i2000SR, ci8200) - Displayed under the right side end cap of the RSH (Example: RSH02001). RSH (i1000SR) - Uses the same 5 digit serial number as the processing module (Example: RSHxxxxx). RSH (c4000, ci4100) - Uses the same 5 digit serial number as the c4000 processing module (Example: RSH01065). Standard Sample Handler (i2000) - Created by the SCC Vendor (the Architect System No. preceded by the letters SH (Example: SH02001). 4. PM Serial Number: c4000 - c4000 for production systems (the ARCHITECT PM serial number preceded by c4; Example: c401065). c8000 - c8xxxxx for production systems (the ARCHITECT PM serial number preceded by c8; Example: c801065). i1000SR - i1SRxxxxx for production systems (the ARCHITECT PM serial number preceded by i1SR; Example: i1SR01065). i2000 - i2xxxxx for production systems (the ARCHITECT PM serial number preceded by i2; Example: i201065). SCC 15 min Reference i2000SR - iSRxxxxx for production systems (the ARCHITECT PM serial number preceded by iSR; Example: iSR01065). Note For multi-module systems, module 1 is on the left. Log On as FSE Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. Note Enter the date code in reverse order. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Architect System Number and SCC Serial Number Configuration 1. From the Snapshot screen, select the System tab. 2. Select Configuration. 3. Select the System Settings radio button. 4. Select the System Control Center category. 5. Select Configure F6. 6. Enter the following serial numbers: Architect system no. Note If replacing the SCC, do not edit the Architect system no. This number remains the same for the system as long as it is at the current site. SCC serial no. Note The system number and SCC serial number are located on the SCC (Platform D - on the front; Platform B or C located on the top of the case). The SCC Serial no. field only accepts eleven charaters. Do not use the alpha charcter (D) that precedes the eleven character numeric SCC serial number. 7. Select Done. Note After editing the serial number, shutdown and restart the SCC to transfer the edits to the database. Sample Handler Serial Note This activity is not required for LAS configurations. Number Configuration 1. From the Snapshot screen, select the System tab. 2. Select Configuration from the dropdown menu. 3. Select the System settings radio button. 4. Select the Sample Handler category. 5. Select Configure F6. 6. Enter the sample handler serial number. 7. Select Done. Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Note After editing the serial number, shutdown and restart the SCC to transfer the edits to the database. Processing Module Serial Number Configuration 1. Select the System icon. 2. Select Configuration from the dropdown menu. 3. Select the System Settings radio button. 4. Select the Modules category. 5. Select Configure F6. 6. Select the desired processing module from the dropdown menu. 7. Enter the processing module serial number. 8. Select Done. 9. After editing the serial number, shutdown and restart the SCC to transfer the edits to the database. Note Do not go to the next step until the system control center displays the Snapshot screen. Enter Serial Numbers in Call Management 1. Enter the serial numbers for the PM(s), SCC, and SH into call management. 2. Enter the Architect System No. into Call Management as the location code for each of the serial numbers. 3. Document the installation per approved country procedures. i System Service and Support Manual (Version 96661-116) • © 2002, 2009 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. c4000, c4100, i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-244 WINMSD Report Version - 96756-111_218_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-244 WINMSD Report Purpose This procedure recalls Windows NT historical information that shows whether NT device drivers have Module SCC been reconfigured since they were set up. Materials Required Floppy or modem for retrieval Action Access System Diagnostics Window Time Steps 1. Select Task Manager from the drop-down menu. 2. Select New Task. 3. The Create New Task dialog displays. 4. Type winmsd in the Open: field and select OK. The Windows NT Diagnostics window displays Create and Save the System Information File 1. Select File, Save Report. The Create Report dialog displays. 2. Select the radio buttons for All Tabs (Scope), Complete (Detail Level), and File (Destination). 3. Select OK. The Save WinMSD Report dialog displays. 4. Leave msdrpt.txt as the File name and select the D: drive at the root level (no subdirectory) as the location. 5. Select Save. The file is save to the D: drive for us to retrieve when drive is sent back. Reference 5 min 6. Close the Windows NT Diagnostics window by clicking on the X in the upper right corner of the window. 7. Close Task Manager by clicking on the X in the upper right corner of the window. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-245 Stop Reagent Onboard Stability Timer Version - 96756-111_219_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-245 Stop Reagent Onboard Stability Timer Purpose This procedure documents how to stop the onboard reagent stability timer. The procedure is only required when: Module SCC a repair significantly affects the time remaining for onboard stability a hardware failure prevents a reagent scan from being performed Note Normally the onboard stability timer is stopped by removing the reagent kits and performing a reagent scan (for the i2000/i2000SR). Materials Required None Time Action Determine time remaining for onboard stability Steps 10 min Reference 1. Select the Reagent icon from the Snapshot screen. 2. Select Reagent status from the drop down menu. 3. Select the View All radio button. 4. Print a listing of all the reagents onboard by pressing ALT+Print Screen. 5. Review the print screen copies with the customer and determine which reagent kits to remove from the module. Log On as FSE Note 1. Select Log on F2 from the Snapshot screen. 2. Enter FSE in all capital letters in the User name field. 3. Enter the FSE logon date code in the password field. 4. Select Done. 5. Verify that the Operator listed on the Snapshot screen lists FSE. Remove Reagent Kits from Module 1. Remove all the reagent kits from the module. If i2000/i2000SR, go to Remove Reagent Kits from Module (i2000/i2000SR). If i1000SR, go to Remove Reagent Kits from Prior to leaving the site, FSE is required to return the system to operator level of access. To do this, perform Log Off as FSE, or restart the SCC. Module (i1000SR). Remove Reagent Kits from Module (i2000/i2000SR) 1. Remove all the reagent kits from the module and store them in the refrigerator. Note Be sure to store all the bottles for each reagent kit together so that there will not be any loading errors when the reagents are loaded on the module again. 2. Go to Stop Reagent Onboard Stability Timer. Remove Reagent Kits from Module (i1000SR) Stop Reagent Onboard Stability Timer 1. Select the Reagent icon from the Snapshot screen. 2. Select Reagent status from the drop down menu. 3. Select the reagent kits to be removed. 4. Select F7-Unload. 1. Stop the reagent onboard stability timer. If i2000/i2000SR, go to Stop Reagent Onboard Stability Timer (i2000/i2000SR). If i1000SR, go to Stop Reagent Onboard Stability Timer (i1000SR). Stop Reagent Onboard Stability Timer (i2000/i2000SR) Note The Remove - F8 function key is only available at the FSE access level. 1. Select the Reagent icon from the Snapshot screen. 2. Select Reagent status from the drop down menu. 3. Select the reagent kits to be removed. 4. Select Remove - F8. 5. Verify removal of the selected reagent kits selected from the reagent status screen. Stop Reagent Onboard Stability Timer (i1000SR) Note The onboard stablity tracking timer stops when the reagent carrier is unloaded from the reagent carousel and removed from the RSH. 1. Select the Reagent icon from the Snapshot screen. 2. Select Reagent status from the drop down menu. 3. Select the reagent kits to be removed. 4. Select Remove - F8. 5. Verify removal of the selected reagent kits selected from the reagent status screen. Note The Remove - F8 function key is only available at the FSE access level. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-248 Startup and Shutdown Version - 96756-111_234_3 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-248 Startup and Shutdown Purpose Identify the recommended steps for a successful startup and shutdown. Module SCC Materials Required None Time 15 min Startup Action Power ON UPS and Peripherals Steps Reference 1. Power ON the UPS. 2. Power ON the peripherals (modem, printer, monitor, etc.). 3. Verify that power to the processing module(s) is OFF. Note The power to the processing module(s) must be OFF prior to turning on the power to the system control center. If the processing module(s) power is ON, communication is not properly initialized between the system components. Power ON SCC 1. Power ON the system control center. Note Do not go to the next step until the system control center displays the Snapshot screen. Power ON Modules 1. Power ON the processing module(s). 2. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Shutdown Action Steps Prerequisite 1. The processing module must be in Stopped, Ready, or Offline status. Perform SCC Shutdown 1. Access the Snapshot screen by selecting Overview and then Snapshot. 2. Select Shutdown F3. Reference 3. Select OK to confirm the Shutdown request. 4. When the screen turns blue, and a message about shutting down the SCC displays, press the Ctrl+Alt+Delete keys simultaneously. Select Shutdown the computer, then OK. Note Wait while the software applications shut down. A message displays informing you that Shutdown is complete and you can turn off the power. Note You can choose to restart the SCC at this point. 5. Press and hold the power switch on the front of the central processing unit (CPU) for 10 seconds to power OFF the SCC. Shut Down Processing Module(s) and Peripherals 1. Power OFF the processing module(s) and all peripherals. i System Service and Support Manual (Version 96661-116) • © 2002, 2008 • ARCHITECT, i1000SR, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, c16000, ci16200, iSystem and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-601 Reagent Supply Center Temperature Detector Adjustment Version - 204739-101_4326_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-601 Reagent Supply Center Temperature Detector Adjustment Purpose Perform this procedure to adjust the displayed temperature of reagent supply center. Module c4000 Materials Required External thermometer (LN8C9488) Insulated cup Ice Time Procedure: 30 min Verification: 30 min Total: 1 hr WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. 3. Remove Temperature Detector from Reagent Supply Center 1. 2. 3. Inspect tools for damage, ensure calibration is not expired and replace if necessary. Verify the processing module has been on for at least 30 minutes prior to testing. The processing module and RSH must be in Stopped status. Loosen the two (2) screws securing the left rear cover and remove the cover. Remove the foam insulation [1] covering the Reagent Supply Center temperature detector. Remove the two (2) screws [2] securing the detector bracket and slide out the detector [3]. Note Do not disconnect the temperature detector from the connector. Determine Water Temperature 1. 2. 3. 4. Fill an insulated cup half full of water. Attach the removed temperature detector to the external thermometer probe with a rubber band and place the probe and sensor into the water. Add ice to cool the water to 5°C ± 3°C. Note This is the testing temperature for this detector. Wait for the reading to stabilize. Determine Reagent 1. Supply Center Temperature 2. 3. Select the As Needed category from the Maintenance screen under the system icon. Perform 3525 Temperature Status. Read the temperature status of reagent supply center. Verify Correct Temperature Reading 1. Verify the reagent supply center temperature is within ± 0.2°C of the reading of the external thermometer in the water. • If correct, go to Reinstall Temperature Detector Into Reagent Supply Center. • If incorrect, continue with Open Card Cage Cover. Open Card Cage Cover 1. 2. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the four (4) thumbscrews securing the card cage cover and open the cover. 3. For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. Adjust VR1 on AC/DC Controller Board 1. 2. Close Card Cage Cover 1. 2. 3. Reinstall Temperature Detector into Reagent Supply Center 1. 2. 3. 4. 5. On the AC/DC controller board, turn VR1 (potentiometer) clockwise to raise the displayed temperature. Note Update the screen after each movement of VR1 to see the current reading. Continue adjustment until the displayed temperature is within ± 0.2°C of the actual temperature. Close the card cage cover and secure with the four (4) thumbscrews. Install the right front access panel and secure with the four (4) screws. Close the right front door. Remove the temperature probe and temperature detector from the ice water and remove rubber band. Remove and dispose of the insulated cup. Gently insert the temperature detector into the reagent supply center and secure with the bracket and two (2) screws. Install the foam insulation covering the temperature detector. Install the left rear cover and secure with the two (2) screws. Prepare for Operation 1. Change the status of the processing module and sample handler from Stopped to Ready by performing the startup procedure. Verification 1. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. For information on the shutdown procedure, refer to the ARCHITECT System Operations Manual, Section 5: Operating instructions, Subsection: System startup, pause, and shutdown. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-602 SCC Bar Code Scanner Verification Version - 204739-101_4327_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-602 SCC Bar Code Scanner Verification Purpose This procedure verifies bar code scanner function. Module Bar Code Scanner Materials Required Bar Code Labels Time 3 min Caution Class 2 Laser Radiation Action Steps Verify SCC BCR Functionality 1. 2. 3. 4. 5. Verification 1. Reference From the Snapshot screen, select the Orders icon. From the drop-down menu, select Patient orders. Select the SID field. Place the SCC bar code scanner over the bar code label. Verify the number from the label displays in the SID field. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-604 Internal Instrument Temperature Detector Adjustment Version - 204739-101_4328_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-604 Internal Instrument Temperature Detector Adjustment Purpose Perform this procedure to adjust the displayed internal c4000 processing module Temperature. Module c4000 Materials Required Standard Tool Kit External thermometer (LN8C94-88) 50mL beaker Time 60 min Caution Possible Electrostatic Discharge Shock Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. 3. Remove Temperature Detector for Water Bath Heater Assembly 1. 2. 3. Determine Water Temperature 1. 2. 3. Determine Internal c4000 Processing Module Temperature 1. 2. 3. Inspect tools for damage, ensure calibration is not expired and replace if necessary. Verify the processing module has been ON for at least 30 minutes prior to testing. The processing module must be in Stopped or Ready status. Loosen the two (2) screws securing the left rear cover and remove the cover. Loosen the screw on the clamp [1] securing the detector [2] at the base of the heater assembly and slide the detector to the left to remove. Do not disconnect the temperature detector from the connector. Fill a beaker with 50mL of room temperature water. Attach the removed temperature detector to the external thermometer probe with a rubber band and place the probes into the water. Wait for the readings to stabilize. Select As Needed category from the Maintenance screen under the System icon. Perform 3525 Temperature Status. Read the status of the instrument internal temperature. For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. Compare Temperature Readings 1. Internal Instrument temperature must be within ±0.2°C of the reading of the external thermometer in the water. • If correct, go to Install Temperature Detector for Water Bath Heater Assembly • If incorrect, continue with Open Card Cage Cover Open Card Cage Cover 1. 2. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the four (4) thumbscrews securing the card cage cover and open the cover. 3. Adjust VR4 1. 2. Close Card Cage Cover 1. 2. 3. Install Temperature Detector for Water Bath Heater Assembly 1. 2. 3. 4. Verification 1. On the AC/DC controller board, turn VR4 (potentiometer) clockwise to raise the displayed temperature and counterclockwise to lower the displayed temperature. Note Update the screen after each movement of the potentiometer to see the current reading. Continue adjustment until the displayed temperature is within ±0.5°C of the actual temperature. Close the card cage cover and secure with the four (4) thumbscrews. Install the right front access panel and secure with the four (4) screws. Close the right front door. Remove probes from the water. Remove rubber band from probes. Slide the detector into the clamp on the base of the heater assembly and tighten the clamp screw to secure the detector. Install the left rear cover. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-605 Water Bath Temperature Detector Adjustment Version - 204739-101_4329_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-605 Water Bath Temperature Detector Adjustment Purpose Perform this procedure to adjust the displayed Water Bath Temperature. Module c4000 Materials Required Standard Tool Kit External thermometer (LN8C94-88) Time 60 min WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Action Steps Prerequisite 1. 2. 3. Reference Inspect tools for damage, ensure calibration is not expired and replace if necessary. Verify the processing module has been ON for at least 30 minutes prior to testing. The processing module must be in Stopped or Ready status. Perform 3526 1. Check Water Bath Temperature 2. Select the As Needed category from the For further information on performing diagnostic procedures, refer to Maintenance screen under the System the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. icon. Perform 3526 Check Water Bath Temperature. Note Allows for correct positioning of the temperature probe in the water bath. Position Probe in Water Bath Position the probe from the external thermometer into the water bath in the position specified by 3526 Check Water Bath Temperature. Wait at least 1 minute for readings to stabilize. 1. 2. Read status of Water Bath Temperature 1. 2. Using 3526 Check Water Bath Temperature read the status of the water bath temperature. The reading should be within ±0.1°C of the reading on the external thermometer. • If correct, go to Prepare for Operation • If the difference is more than ±0.1°C go to Open Card Cage Cover Open Card Cage Cover 1. 2. 3. Adjust VR5 1. 2. Close Card Cage Cover 1. 2. 3. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the four (4) thumbscrews securing the card cage cover and open the cover. On the AC/DC controller board, turn VR5 (potentiometer) clockwise to raise the displayed temperature and counterclockwise to lower the displayed temperature. Note Update the screen after each movement of the potentiometer to see the current reading. Continue adjustment until the displayed temperature is within ±0.1°C of the actual temperature. Close the card cage cover and secure with the four (4) thumbscrews. Install the right front access panel and secure with the four (4) screws. Close the right front door. Prepare for Operation 1. Complete 3526 Check Water Bath Temperature and remove the external thermometer probe. Verification 1. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-606 Reagent Supply Center Bar Code Reader Alignment Version - 204739-101_4330_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-606 Reagent Supply Center Bar Code Reader/Bottle Sensor Alignment Purpose Perform this procedure to align the Reagent Supply Center Bar Code Reader and Bottle Sensor. Module c4000 Materials Required Standard Tool Kit PM10-63690*AGr1 Bar Code Reader JIG tool 2-202076-01 JIGU Cable ISA ARCH-011 Updated JADAK 2D Barcode Reader Command Card Time 20 min WARNING Potential Biohazard Caution Chemical Hazard Action Steps Prerequisite 1. 2. 3. Preparation 1. 2. 3. 4. 5. Inspect tools for damage, ensure calibration is not expired and replace if necessary. The processing module must be in Stopped or Ready status. Log On as FSE. Open the Reagent Supply Center cover and remove any reagents in segment J. Store in appropriate refrigerator. Remove the left rear cover. Remove the left side cover to provide access to the bar code reader adjustment screws. [1] Remove the two (2) screws securing the front left deck cover [A]. Position the cover so there is access to the bar code reader. Do not disconnect the wires to the reagent carousel cover and rotation buttons. Reference Preparation (continued) 1. 2. Adjust the red circle screw to the uppermost position of the oblong screw hole. Adjust the blue circle screws to the center of the screw holes. Move The Reagent Carousel To Home Position 1. 2. 3. Put Bar Code Reader in “Lights Off” Aimer Mode 1. 2. 3. 4. From the snapshot screen, select System, Diagnostics, Module 5, Utilities and perform 6200 CLI Terminal Simulator. Enter set_module 1 continue, r1oc h bcr continue, and r1ic h bcr continue to home the inner and outer carousels. Verify the mark and the slit are aligned straight. Disconnect the barcode reader cable from the reader and the D-sub connector. Connect the JIGU cable between the Dsub connector and the barcode reader. Note Ensure the switch on the JIGU cable is OFF.[1] Turn on the JIGU cable switch. The barcode reader is triggered on and the red illumination and the green aimer light up. Position the Lights Off command barcode in front of the barcode reader (Inside the Reagent Supply Center behind the window). 5. Turn off the switch on the JIGU cable. Then turn the switch on the JIGU cable back on. Only the green aimer lights up. Put Bar Code Reader in “Lights Off” Aimer Mode (continued) Install JIG Alignment Tool Per ISA ARCH-011 Updated JADAK 2D Barcode Reader Command Card. 1. Install the JIG Alignment Tool into the BCR tray.[1] Note The edge of the JIG tool must fit into the right corner of the BCRTRSAY. Adjust BCR Position 1. 2. Locate the BCR angle, vertical and horizontal adjustment screws (C, D & E). Adjust the BCR position so the center of the Aimer Light (green) is positioned on both the front and back of the target. Put Bar Code Reader in “Lights On” Mode 1. 2. 3. Adjust Bottle 1. Sensor Position 2. 3. 4. Bottle Sensor Sensitivity Adjustment 1. Scan the Lights On barcode label at the BCR Tray. Turn off the switch on the JIGU cable. Then turn the switch on the JIGU cable back on. Verify the Illumination Light (red) is lit. Remove the JIGU cable and reconnect the bar code reader cable. Using 6200 CLI Terminal Simulator, type C8000_COMMAND <R1IC H BSPOS> to move the carousel to the bottle sensor adjustment position. Locate the bottle sensor vertical and horizontal adjustment screws (A &B). Adjust the Bottle Sensor position so the red beam hits the back target plate through the hole on the front plate. Remove the JIG Alignment Tool. Perform 6200 CLI Terminal Simulator and type r1ic pos bs k 1 to move the first position of segment K to the Bottle Sensor adjustment position. a. Verify the Mode Switch on the Bottle Sensor is turned clockwise fully. b. Place a 90ml reagent cartridge into the first position of the K segment. c. Turn the sensitivity volume CCW fully for the 2. 3. 4. 5. 6. 7. 8. Verification Read Test 1. 2. 3. Prepare for Operation 1. Ensure reagent cartridges are loaded on the Reagent Carousel. From the Snapshot screen, select Reagents, Reagent Status and Scan. Verify all regents were read correctly and displayed in the Reagent Status screen. 3. Install the front left deck cover. Install the left rear and left side covers. Log On as Operator. 1. Perform System Start- 2. Verification lowest sensitivity. d. Turn the sensitivity volume CW gradually until the Active LED (red) is lit. Record the position as “A”. e. Remove the 90ml reagent cartridge. Turn the sensitivity volume CW gradually until the Active LED (red) is lit and then turn the volume CCW gradually until the Active LED goes OFF. Record the position as “B”. Note If the red light is not lit at the maximum sensitivity point, record the position as “B”. Set the sensitivity volume at the intermediate position of A and B. Verify Active LED (red) is OFF. Place a 90mL reagent cartridge into the first position of K segment and verify the Active LED (red) is lit. Remove the 90mL reagent cartridge. Type R1ic h and r1oc h to home the inner and outer carousels. Exit 6200 CLI Terminal Simulator. 2. up to put the processing module into the Ready Status. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2013 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-608 Sample Wash Cup Position Alignment Version - 204739-101_4331_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-608 Sample Wash Cup Position Alignment Purpose Perform this procedure to align (mechanically) the Sample Wash Cup to Sample Probe. Module c4000 Materials Required Standard Tool Kit Time 15 min WARNING Potential Biohazard WARNING Probe Stick Hazard Action Steps Prerequisite 1. 2. Inspect tools for damage, ensure For further information on performing diagnostic procedures, refer to calibration is not expired and replace if the ARCHITECT System Operations Manual, Section 10: necessary. Troubleshooting and diagnostics, Subsection: System diagnostics. Power OFF the c4000 processing module. Verify Sample Pipettor Sample Wash Cup 1. Manually move the sample pipettor through the sample wash cup to ensure that probe passes through the cup without contacting the cup. Note Ensure probe is new and straight. Power On 1. 2. Power on the c4000 processing module. The processing module must be in Stopped or Ready status. Log On as FSE. 3. Perform 1124 Wash Cup Alignment 1. 2. 3. 4. Adjust Sample Wash Cup 1. 2. Perform 1120 Sample Pipettor Calibration 1. 2. Reference Select the Fluidics / Wash category from the Diagnostics screen under the System icon. Perform 1124 Wash Cup Alignment. Note Allows for correct positioning of the sample probe for alignment of the wash cup. Select L1 for the Sample Pipettor. Select L1 to home the pipettor over the wash cup. Loosen the screws on the base of the sample wash cup and center the wash cup under the sample probe. Note Ensure the probe does not contact the wash cup. Tighten the screws on base of sample wash cup once wash cup is in position. Select the As Needed category from the For further information on performing diagnostic procedures, refer to Maintenance screen under the System the ARCHITECT System Operations Manual, Section 10: icon. Troubleshooting and diagnostics, Subsection: System diagnostics. Perform 1120 Sample Pipettor Calibration. Verification 3. Log On as Operator. 1. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-609 Degasser Canister Draining Version - 204739-101_4332_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-609 Degasser Canister Draining Purpose Perform this procedure to remove fluid from the degasser unit canister. Module c4000 Materials Required Standard Tool Kit Syringe 30 min Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. Preparation 1. 2. Drain Water from Canister 1. 2. 3. 4. 5. Prepare for Operation 1. 2. 3. 4. 5. 6. Verification 1. Inspect tools for damage, ensure calibration is not expired and replace if necessary. The processing module must be in Stopped or Offline status. Time Power OFF the c4000 processing module (use the main circuit breaker located at the rear of the processing module). Remove the left rear cover. Open the left rear cover. Locate canister drain tubing in front of degasser and remove from clamp. Remove cap and attach a syringe to the end of the drain tubing. Draw back on the syringe plunger to remove water from the degasser canister. Repeat as needed to remove all water from the degasser canister and install cap on drain tubing. Power ON the processing module. Note The system control center power must be ON prior to turning on the processing module to ensure proper initialization. To change the status of the processing module and sample handler from Stopped to Ready, perform the Startup procedure. Select the As Needed category from the Maintenance screen under the system icon. Perform 2132 Flush Water Lines. Check for leaks. Install the left rear cover. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-610 Reagent Wash Cup (R1 and R2) Alignment Version - 204739-101_4333_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-610 Reagent Wash Cup (R1 and R2) Alignment Purpose Perform this procedure to align (mechanically) the Reagent Wash Cups to Reagent Probes. Module c4000 Materials Required Standard Tool Kit Time 15 min WARNING Potential Biohazard Caution Chemical Hazard WARNING Action Steps Prerequisite 1. 2. Probe Stick Hazard Inspect tools for damage, ensure For further information on performing diagnostic procedures, refer to calibration is not expired and replace if the ARCHITECT System Operations Manual, Section 10: necessary. Troubleshooting and diagnostics, Subsection: System diagnostics. Power OFF the c4000 processing module. Verify Reagent Pipettor Does Not Contact Reagent Wash Cup 1. Manually move the reagent pipettor through the reagent wash cup to ensure the probe passes through the cup without contacting the cup. Note Ensure probe is new and straight. Power on the c4000 1. 2. Power on the c4000 processing module. The processing module must be in Stopped or Ready status. Log On as FSE. 3. Perform 1124 Wash Cup Alignment 1. 2. 3. Adjust Sample Wash Cup 1. 2. 3. Reference Select the Fluidics / Wash category from the Diagnostics screen under the System icon. Perform 1124 Wash Cup Alignment. Note Allows for correct positioning of the reagent probe for alignment of the wash cup. Select L2 (R1) or L3 (R2) for the R1 or R2 pipettors. Loosen the screws on the base of the reagent wash cup and center the cup under the reagent probe. Note Ensure the probe does not contact the wash cup. Adjust as needed. Tighten the screws on the base of the reagent wash cup once the wash cup is in position. Calibrate Pipettor 1. 2. 3. Verification 1. Select the As Needed category from the Maintenance screen under the System icon. Perform 1121 R1 Pipettor Calibration or 1122 R2 Pipettor Calibration. Log On as Operator. For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-611 Wash Cup Position Alignment Version - 204739-101_4334_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-611 Wash Cup Position Alignment Purpose Perform this procedure to align (mechanically) the Mixer Wash Cups to the mixers. Module c4000 Materials Required Standard Tool Kit Time 15 min WARNING Potential Biohazard Caution Chemical Hazard WARNING Probe Stick Hazard Action Steps Reference Prerequisite 1. For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. 2. The processing module must be in Stopped or Ready status. Log On as FSE. Verify Mixer Pipettor Position 1. Manually move the mixer through the mixer cup to ensure the mixer does not contact the mixer wash cup. Perform 1123 Mixer Alignment 1. Select the Fluidics / Wash category from the Diagnostics screen under the System icon. Perform 1123 Mixer Alignment. Note Allows for correct positioning of the mixer for alignment of the wash cup. Select L1 for Mixer 1 or L2 for Mixer. Select L2 to position the mixer over the Wash Cup. 2. 3. 4. Adjust Mixer Cup 1. 2. 3. 4. Verification 1. Loosen the screws at the base of the wash cup and position the wash cup so that it is centered under the mixer. Note Ensure the mixers do not contact the wash cup. Once the cup is in the correct position, tighten the screws at the base of the wash cup. When adjustment is correct, exit 1123 Mixer Alignment. Log On as Operator. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-612 Bulk Solution Weight Sensor Adjustment Version - 204739-101_4335_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-612 Bulk Solution Weight Sensor Adjustment Purpose Perform this procedure to adjust the bulk solution weigh sensors for minimum volume in the bulk solutions after a repair or replacement. Module c4000 Materials Required Standard Tool Kit 1 empty ICT Reference (large) bottle 1 empty alkaline or acid (small) bottle 1 metric ruler Absorbent Towels Time 30 min WARNING Potential Biohazard Caution Chemical Hazard Caution Possible Electrostatic Discharge Shock Action Steps Preparation 1. 2. 3. 4. 5. Fill Bulk Solution Bottles 1. 2. Open Card Cage Cover 1. 2. 3. Access MECH LED Board 1. Inspect tools for damage, ensure calibration is not expired and replace if necessary. The processing module must be in Stopped or Ready status. Open the supply (left front) door. Remove the tubing from the ICT reference bottle, alkaline wash bottle, and acid wash bottle. Wrap each tube with an absorbent towel. Remove the ICT reference bottle, alkaline wash bottle, and acid wash bottle from the weight sensor platforms. Fill the empty ICT reference solution bottle with 15mm of water (165mL +/10mL). Fill the empty alkaline or acid bottle with 14mm of water (67mL +/- 5mL). Open the right front door. Remove the right front panel by removing the four (4) screws. Loosen the four (4) thumbscrews securing the card cage cover and open the cover. Verify the MECH LED board [1] is visible from the left side of the Reference processing module. Position Bottles on Platform and Verify LEDs are ON 1. 2. Verify Weight Sensor Specification 1. 2. 3. Adjust Weight Sensors 1. 2. Position the test bottle on the appropriate platform. Note Do not place intake tubing in the bottle. Verify the LED on the MECH LED Board is ON. • ICT reference = LED 9-7 Empty Sensor MCC Warning • Alkaline = LED 9-8 Empty Sensor B14 Warning • Acid = LED 9-9 Empty Sensor B3 Warning Pipette 11mL of water from the empty ICT reference solution bottle. This should lower the height of the water by 1mm (14mm). Pipette 5mL of water from the empty alkaline or acid bottle. This should lower the height of the water by 1mm (13mm). Verify the LED on the MECH LED board is OFF. Note The specification is for the ICT reference solution LED to be on for 15mm and off at 14mm. The LED for alkaline or acid to be on at 14mm and off 13mm. • If a weight sensor is in specification, go to Prepare for Operation. • If a weight sensor is not in specification, go to Adjust Weight Sensors. Loosen the nut [1] on the flag adjustment and turn the screw [2] CW if the LED is OFF at the upper water volume or CCW if the LED is ON at the lower water volume. Tighten nut [1] and retest. Prepare for Operation 1. 2. 3. 4. 5. Verification 1. Install the ICT reference bottle, alkaline wash bottle, and acid wash bottle on the weight sensor platforms. Install the tubing in the ICT reference bottle, alkaline wash bottle, and acid wash bottle. Replace all cover(s) and close all doors. Select the As Needed category from the Maintenance screen under the System icon. Perform 2155 Flush Bulk Solutions. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-613 +5VDC (PS1) Measurement Version - 204739-101_4336_2 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-613 +5VDC (PS1) Measurement Purpose Perform this procedure to measure +5VDC (PS1) in the Main Power Supply. Module c4000 Materials Required Standard Tool Kit Multimeter Time 45 min WARNING Potential Biohazard WARNING Electrical Shock Hazard Caution Possible Electrostatic Discharge Shock Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. 3. Open Card Cage Cover 1. 2. 3. Measure +5VDC on DAQ Board 1. Inspect tools for damage, ensure calibration is not expired and replace if necessary. The processing module must be in Stopped or Ready status. Log On as FSE. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the four (4) thumbscrews securing the card cage front cover and open the cover. Verify the voltage between TP3 (+ lead) and TP4 (- lead) on the DAQ board is +5.05VDC (±0.05). • If the voltage is correct go to Close Card Cage Cover. • If the voltage is incorrect, proceed to action item Adjust Voltage. Adjust Voltage 1. 2. 3. 4. Close Card Cage Cover 1. 2. 3. Remove the right rear cover. Locate the variable resistors [1] on the back of the Main Power Supply. Adjust the variable resistor for PS1 for a voltage reading of +5.05VDC (±0.05) between TP3 (+ lead) and TP4 (- lead) on the DAQ board. Install the right rear cover. Close the card cage cover and secure with the four (4) thumbscrews. Install the right front access panel and secure with the four (4) screws. Close the right front door. Prepare for Operation 1. To change the status of the processing For information on the startup procedure, refer to the ARCHITECT module and sample handler from System Operations Manual, Section 5: Operating instructions, Stopped to Ready, perform the Startup Subsection: System startup, pause, and shutdown. procedure. Verification 1. Select the Fuses / Motors category For further information on performing diagnostic procedures, refer to the from the Diagnostics screen under the ARCHITECT System Operations Manual, Section 10: Troubleshooting System icon. and diagnostics, Subsection: System diagnostics. Perform 5712 Voltages Test. Log On as Operator. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. 2. 3. 4. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-614 +12VDC (PS3 Power Supply) Measurement Version - 204739-101_4337_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-614 +12VDC (PS3 Power Supply) Measurement Purpose Perform this procedure to measure the +12 VDC on the AC/DC Driver Board. Module c4000 Materials Required Standard Tool Kit Multimeter Time 45 min WARNING Potential Biohazard WARNING Electrical Shock Hazard Caution Possible Electrostatic Discharge Shock Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. Open Card Cage Cover 1. 2. 3. 4. Inspect tools for damage, ensure calibration is not expired and replace if necessary. The processing module must be in Stopped or Ready status. Open the right front door. Remove the right front access panel by removing the four (4) screws. Loosen the two (2) screws on the Driver Box door [1] and open the door. Locate CN12 [2] on the AC/DC Driver Board. Measure +12 VDC on AC/DC Driver Board 1. Verify that voltage between Pin 1 (+ lead) and Pin 2 (- lead) on CN12 is +12VDC (±0.3). • If the voltage is correct, proceed to action item Close Front Covers. • If the voltage is incorrect, proceed to action item Adjust Voltage. Adjust Voltage 1. 2. 3. Remove the right rear cover. Locate the variable resistors [1] on the back of the Main Power Supply. Adjust the variable resistor for PS3 for a voltage reading of +12VDC (±0.3) between Pin 1 (+ lead) and Pin 2 (- lead) on CN12 on the AC/DC Driver Board. Install Right Rear Cover 1. Install the right rear cover. Close Front Covers 1. Close the Driver Box door and secure with the two (2) screws. Install the right front access panel and secure with the four (4) screws. Close the right front door. 2. 3. Verification 1. 2. Select the Fuses / Motors category from the Diagnostics screen under the System icon. Perform 5712 Voltages Test. 3. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. Note For further information on performing diagnostic procedures, refer to the ARCHITECT System Operations Manual, Section 10: Troubleshooting and diagnostics, Subsection: System diagnostics. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-615 +12VDC (PS5) Measurement Version - 204739-101_4338_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-615 +12VDC (PS5) Measurement Purpose Perform this procedure to measure +12VDC (PS5) in the Main Power Supply. Module c4000 Materials Required Standard Tool Kit Multimeter Time 45 min WARNING Potential Biohazard WARNING Electrical Shock Hazard Caution Possible Electrostatic Discharge Shock Action Steps Reference Prerequisite 1. For further information, refer to the ARCHITECT System Operations Manual, Section 1: Use or function, Subsection: System status. 2. Inspect tools for damage, ensure calibration is not expired and replace if necessary. The processing module must be in Stopped or Ready status. Access Main Power Supply 1. Remove the right rear cover to access the Main Power Supply and power connectors. [1] Verify +12VDC (PS5) at Unit Cooler Plug 1. Locate the UC connector [1] located on the side of the Main Power Supply. Use a digital voltmeter to verify the voltage between pin 1 (+ lead, orange wire) and pin 2 (- lead, black wire) is +12VDC (±0.6VDC). Note This voltage is only present 2. when the reagent carousel is cooling. • If the voltage is correct, go to Install Right Rear Cover. • If the voltage is incorrect, proceed to action item Adjust Voltage. Note UC Connector 123 456 789 Adjust Voltage 1. 2. Locate the variable resistors [1] on the Main Power Supply. Adjust the PS5 +12V variable resistor for a voltage reading of +12VDC (±0.6VDC) between pin 1 (+ lead, orange wire) and pin 2 (- lead, black wire) is on the UC connector. Install Right Rear Cover 1. Install the right rear cover. Verification 1. Select the Fuses / Motors category For further information on performing diagnostic procedures, refer to from the Diagnostics screen under the the ARCHITECT System Operations Manual, Section 10: System icon. Troubleshooting and diagnostics, Subsection: System diagnostics. Perform 5712 Voltages Test. Log On as Operator. After repair is complete, ensure appropriate Quality Control is in specification and calibrate as necessary. 2. 3. 4. ARCHITECT c4000 Service and Support Manual (Version 204733-102) • © 2009 • ARCHITECT, i1000SR, i2000, i2000SR, c8000, ci8200 are registered trademarks of Abbott Laboratories. c4000, ci4100, i4000, i6000, i8000, c16000, ci16200 and ARCHITECT ARM are trademarks of Abbott Laboratories. • Abbott Park, IL 60064 • All rights reserved. P-616 +11.45 VDC, Lamp Power Supply (PS4) Measurement Version - 204739-101_4339_1 Inspect tools for damage, ensure calibration is not expired and replace if necessary. At the completion of this VP: G110 - After repair is complete, verify per released Operation and Service Procedures. If the system/instrument produces results, ENSURE appropriate Quality Control is in specification and calibrate as necessary. P-616 +11.45 VDC, Lamp Power Supply (PS4) Measurement Purpose Perform this procedure to adjust DC Lamp Power Supply in the Main Power Supply. Module c4000 Materials Re
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )