LGI injection rate Auteurs S. Winkelman Address a DaVinci Laboratory Solutions Sydneystraat 5, 3047 BP Rotterdam ARTIKEL INFO ABSTRACT Keywords: Internal document for the LGIvolumetric injection rate Davinci Laboratory Solutions uses an automotive injection valve integrated with a needle mounted on a chromatograph (GC) to inject high pressure liquefied gas named: LGI (Liquefied gas injection). In this paper two experiments are described. One focusses on the effects of volume injection with GC system back pressure. By injecting into a closed system pressurized by nitrogen a backpressure effect is simulated. The results are presented in Table 1: GC backpressure effect and a conclusion can be made that backpressure does not affect the volume injected into the system. The second experiment focusses on the repeatability and reproducibility of the injector/GC needle valve combination. This done by injecting into a vial and weighing the amount injected. The results are presented in Table 2: Repeatability: Volumetric injection LGI & Table 3: Reproducibility: experiments. After applying statistics described in the ASTM D7756 norm it is concluded that: That the injector-needle combination with a 99% certainty a volumetric injection presented in Appendix 4 with a 95% probability that these values interchangeable between injectors. Hereby effectively reporting the repeatability and reproducibility of the product. 1 Introduction The company Davinci Laboratory Solutions (Davinci-LS) has used an automotive injection valve to automatically inject a high pressure sample directly into a gas chromatograph (GC) called LGI (Liquefied gas injection). The system uses a Hitachi FSI injector rated up to a pressure of 80 bar common rail pressure. The injector has been modified to fit a custom GC needle effectively creating a high-pressure injection sample system to inject liquid high pressure samples. This system can used to inject samples form a distillation tower or oil well to measure a representative pressurized process sample in a GC. To laboratory technicians its valuable to know how much and how repeatable the amount injected is. The rate of injection of a FSI injector has been widely modelled in literature [1] [2] [3]. The rate of injection is dependent on the speed of sound in the fluid and the injection orifice explained by J. T. Menucci. The mathematical model is described by the following variables: Common rail pressure, cylinder head pressure, Injector orifice resistance and time [2]: 𝑑(𝑚) 𝐴 = ∗ 𝑃 [2] 𝑑𝑡 𝑎 Where m is in kilogram injected mass, t in milliseconds open/close time, A in m2 as injection orifice, a in m/s as the speed of sound trough the liquefied gas and P in bar as common rail pressure (gas sample bomb pressure). The speed of sound in the formula is used because a sudden mechanical movement of fluid under pressure is governed by the sound of sound through the fluid [1]. This short paper focuses on modelling the injection rate of the FSIcustom needle combination. This will be done by looking at the effects of GC backpressure (cylinder head pressure) and the overall injection rate of the system. The results are statistically analysed for the repeatability and reproducibility plus the mathematical model is validated. 2 Experiments Both experiments are based on the work done by J. T. Menucci and are adapted to fit the application [2]. 2.1 GC backpressure effects To simulate the effects of backpressure on the injection needle the needle must inject into a closed system under a specific GC inlet pressure ranging from 2 – 6 bar absolute. This is done by welding a GC inlet onto a 1/16-inch tubing with a fixed geometry and bending the tube 90 degrees and then filling the fixed geometry with octane so that the injection is done in a fluid which equally divides the pressure in the system. By using a 6-way valve with a small internal volume the system can be pressurized with N2 and closed when the desired pressure is reached. See Appendix 1 for a piping & instrumentation diagram (PID), Custom GC inlet and automotive adapter. When injecting a fluid in a closed system results in a pressure increase and then the injected volume can be calculated by using the ideal gas law: 𝑃1 ∗ 𝑉1 = 𝑃2 ∗ 𝑉2 𝑉𝑔𝑠 ∗ 𝑃0 𝑉𝑔𝑠−𝑛𝑑𝑙𝑒 = 𝑃𝑛𝑙𝑑𝑒 𝑉𝑔𝑠−𝑛𝑑𝑙𝑒 ∗ 𝑃𝑛𝑑𝑙𝑒 𝑉𝑔𝑠−𝑛𝑑𝑙𝑒−𝑖𝑛𝑗 = 𝑃1 𝑉𝑖𝑛𝑗 = 𝑉𝑔𝑠 − 𝑉𝑔𝑠−𝑛𝑑𝑙𝑒 − 𝑉𝑔𝑠−𝑛𝑑𝑙𝑒−𝑖𝑛𝑗 Subsidize and eliminate. 𝑃0 𝑃0 𝑉𝑖𝑛𝑗 = 𝑉𝑔𝑠 ∗ ( − ) 𝑃𝑛𝑑𝑙𝑒 𝑃1 Figure 1: schematic of setup injecting multiple times, the mass will stepwise increase, then by subtracting each value the mass the single injection mass will be found. See Appendix 2 for the PID and code. Materials used: Swagelok Port connectors. Swagelok male female connectors. Swagelok bolts. Swagelok 1/16 inch tubing. Swagelok butterfly valve. Sample Bomb (100 bar). FSI (Hitachi): 06E 906 036AE with GC needle. 100 Bar N2. Octane stock. Injector driver module. Analytical scale. KERN ARS 120-4. Swagelok EN 837-1 PC with C+ code and serial connection to injection module and scale. 2.3 Method used Described in paragraph 2.3.1 and 2.3.1 are the procedure for the experiments. 2.3.1 Materials used: Swagelok Port connectors. Swagelok male female connectors. Swagelok bolts. Swagelok 1/16-inch tubing. Swagelok butterfly valve. Swagelok needle valve. Sample Bomb (100 bar). Omega Pressure sensor Model No: PXM319-200G10V with custom digital readout using a We!ntek MT8050iE Human Machine Interface. FSI (Hitachi): 06E 906 036AE with GC needle. Injector driver module. 100 Bar N2. Octane stock. Septum + custom made GC septum inlet. 6 way valve. 2.2 Injection amount The experiment is done by injecting into an empty vial that is placed on an analytical scale. Code is written in C+ to automatically inject a sample and log the analytical scale value into a .CSV file. By GC backpressure Variable: Pressure 1,2,3,4,6 bar absolute measurement tube & 10 milliseconds injection time & 20, 40, 60, 80 bar Sample Bomb pressure. 1. Fill the Sample Bomb with 40 ml octane. 2. Pressurize the Sample bomb with 100 bar N2 and connect to the system. 3. Open sample bomb valves and spool the system. 4. Open the GC inlet. 5. Fill GC inlet with octane up to the brim and shake air bubbles out of the system. 6. Close GC inlet with septum. 7. Turn 6-way valve into fill configuration. 8. Pressurize the system. 9. Turn the 6-way valve into measurement configuration. 10. Measure pressure. 11. Inject GC needle into GC inlet. 12. Measure pressure 13. Injection. 14. Measure pressure 2.3.2 Overall rate of injection Variable: 25,50 milliseconds injection time & 20, 40, 60, 80 bar Sample Bomb pressure, FSI (Hitachi). 1. Fill the Sample Bomb with 40 ml octane. 2. Pressurize the Sample bomb with 100 bar N2 and connect to the system. 3. Open sample bomb valves and flush the system. 4. Place setup above vial on scale. 5. Connect serial connection to PC. 6. Run code. 7. Depressurize the sample bomb pressure to next value. 8. Repeat till 20 bar. 9. Swap FSI for another one. 3 Result ANOVA statistics 3.1 GC backpressure effect H0 = P(0,05) F F crit P-value dF 20 bar, 50ms 190,7 3,1 0,000 81 40 bar, 50ms 19,2 3,1 0,000 77 60 bar, 50ms 0,0 3,1 0,996 80 80 bar, 50ms 124,4 3,1 0,000 75 20 bar, 25ms 3,1 3,1 0,051 83 40 bar, 25ms 66,9 3,1 0,000 86 60 bar, 25ms 88,2 3,1 0,000 86 80 bar, 25ms 89,6 3,1 0,000 84 After completing the experiment, the data is calculated and plotted in the graph below. The backpressure influences the injection volume expected was that the injection volume decreased if pressure increased but as shown below this is not the case. DATA, Back pressure effect V(uL) 5,0 4,0 20 bar 3,0 40 bar 2,0 60 bar 1,0 80 bar Table 3: Reproducibility: experiments 3.3 Model validation 0,0 0 2 4 6 8 P(bar) relative back pressure Table 1: GC backpressure effect Using the physical values below and experimental values displayed in Appendix 4 a correlation graph is made with an ANOVA analysis with a H0 = P(0.05): Table 4: Physical value's Physical value's After completing the experiments, the data is processed in excel with suggested statistics from the product current ASTM D7756 norm form the LGI-system, see Appendix 3 for the raw data output. After calculating the mass measured per single injection statistics is applied using t-student statistics to calculate the upper and lower limit of injection with a 99% certainty effectively displaying the repeatability of the injector with a GC needle. See Appendix 4 for the actual values. Spread with 99% confidence interval 100 a.octane(m/s) 1171 A.needle(m2) 4,9E-08 model vs experiment 100 Lineair (m.model, 50ms) Lineair (m.experime ntal, 50ms) Lineair (m.model, 25ms) Lineair (m.experime ntal, 25ms) 80 P (bar) 3.2 Overall rate of injection 60 40 p(bar) 20 0 0,0E+00 50 1,0E-05 2,0E-05 mass (kg) Table 5: model vs experiment 0 0,0 5,0 10,0 15,0 50ms, mu1 V(uL) 25ms, mu1 20,0 25,0 50ms, mu2 25ms, mu2 Table 2: Repeatability: Volumetric injection LGI H0 = 0,05 F F crit P-value dF 25 ms 0,29 5,98 0,60 7 50 ms 0,86 5,98 0,38 7 Table 6: model vs experiment The FSI is swapped out 3 times to perform the same experiment where after that the limit values are compared using a single side analysis of variance (ANOVA) statistics effectively displaying the reproducibility of the injector with a GC needle with a H0 =P(0.05) accepted variation. As seen below the hypnotises can be accepted in multiple cases, so the injector with GC needle produces statistically reproducible volume injections. Why the F value is in some cases higher than the F crit value will be discussed in section 4. Looking at the F values there can be concluded that the model has validation and the 50 ms hypothesis is accepted while 25 ms is rejected. 4 Discussion The experiments described in section 3.2 are done by manually venting N2 from 80 bar to 20 bar with a manometer. A discussion can be made about the fact that the pressure was measured analogously and was therefore not exactly at the setpoint ranging from 80 till 20 bar. Therefore, the F values presented in Table 3 can be explained, the injection volume is strongly dependent on the gas sample bomb pressure and the volumes will thus not me the same if the pressure is slightly different. The same explanation can be made for the P values of table 6. 6 Appendix 6.1 Appendix 1 5 Conclusion A system designed by Davinci-LS uses an automotive injector valve. This valve is modified with a gas chromatograph needle to be used as high-pressure liquefied gas injection into a gas chromatograph. An experiment has been setup to measure the effects of system back pressure and injection rate of the system. After reviewing the back pressure experiment there can be concluded that backpressure does not affect the injection volume. After preforming the rate of injection experiment 3 times that by injecting though a custom needle, swapping the injector and performing statistics on the result a conclusion can be made that: with 99% certainty the injected volume octane is as is presented in table 2 and Appendix 4 effectively displaying the repeatability of the product. Then the reproducibility is calculated, aside from value’s diverting due to a gas bomb pressure setpoint deviation (as described in the discussion) the hypothesis is accepted that 95% of an x number of injections lay between the values shown in table 2 and Appendix 4. After comparing the model values with the experimental values a deviation ranging from 50% to 5% is found between the two. The cause of this variation is due to the seen leakage of the injector when the custom needle is attached. Further research is needed why the injector leaks when the needle is attached. 6. Bibliography Figure 2: PID setup 1 [1] W. Bosch, “The Fuel rate Indicator: A new measuring Instrument For Display of the Characteristics of Individual Injection,” SAE International, USA, 1967. [2] T. J. Menucci, “Development of Bosch Rate of Injection measurement Procedure and Results,” Michigan Technological university, Michigan, 2018. [3] M. Shashank, “Piezoelectric Diesel Injectors & Emission Control,” International Journal of Science and Research, India, 2015. [4] Bosch, “HDEV6: Injection pressures up to 350 bar,” 2018. [Online]. Avialable: https://www.bosch-mobility-solutions.com/en/solutions/valves/highpressure-injector/. Figure 3: Custom GC inlet with septum and pressure sensor Figure 4: Injector plus custom needle adapter 6.2 Appendix 2 Figure 6: injection Module, scale serial connection Figure 5: PID setup 2 Figure 7: Experimental setup 6.3 Appendix 3 Table 7: Raw data 50ms injectie 0,5 0,45 0,4 mass(g) 0,35 0,3 0,25 0,2 0,15 0,1 0,05 0 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 injection counter 20bar, 25ms gewicht 40bar, 25ms mass 60bar, 25ms mass 80bar, 25ms mass Table 8: Raw data 25ms injectie 0,25 mass (g) 0,2 0,15 0,1 0,05 0 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 injection counter 20bar, 25ms gewicht 40bar, 25ms mass 60bar, 25ms mass 80bar, 25ms mass 6.4 Appendix 4 t-student V.limits (uL) 20 bar, 50ms 11,0 10,4 40 bar, 50ms 16,8 16,4 60 bar, 50ms 20,5 20,2 80 bar, 50ms 23,4 23,1 20 bar, 25ms 6,4 5,9 40 bar, 25ms 9,6 9,1 60 bar, 25ms 11,3 11,0 80 bar, 25ms 12,8 12,4
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )