ANNEXURE A1.1: TECHNICAL SPECIFICATION
Model: Allengers 30
X-Ray Generator
Output Power
1.3 KW
Max. mA Output
30 mA
KV Range
52-85 KV
mAs Range
0.9-120 mAs
Exposure Time
0.06-8 Sec in 24 steps
mA Selection
15, 20, 30
Collimator
Manual LBD
Input Power Line Requirement
Supply Configuration
Single phase AC Voltage
Input Voltage
230VAC ±10%
Line Frequency
50/60 Hz.
Max. Line Resistance
1.0 Ohms
Main fuse Rating
10 A
Ambient Conditions
Storage Conditions
Temperature
00 - 400C
Relative Humidity
Maximum 90% at 300C
Operating Conditions
Temperature
100 - 400C
Relative Humidity
30 to 75% RH Non Condensing
Atmospheric pressure
70KPa…110KPa