ESSENTIAL RANGE RIOS/ELIX/AFS ELECTRONICS & TROUBLESHOOTING GUIDE SERVICE MANUAL SM_ESSENTIAL RANGE_TC&EC Safety information Safety symbols Symbol ! What it means This HAZARD symbol is used to refer to instructions in this manual that need to be done safely and carefully. This ATTENTION symbol is used to refer to instructions in this manual that need to be done carefully. This UV RADIATION sticker is used to refer to a position on the Water System Cabinet or inside of it where exposure to UV light is possible. This DANGER sticker is used to refer to a position on the Water System Cabinet or inside of it that could be hazardous. This ELECTRICAL GROUND sticker is used to refer to a position on the Water System Cabinet or inside where an electrical ground connection is made. This ELECTRICAL DANGER sticker is used to refer to a position on the Water System Cabinet or inside where an electrical danger could exist. Important information ! HAZARD ! ATTENTION SM_ESSENTIAL RANGE_TC&EC The Water System must be connected to a source of electrical power that is Earth Grounded. When the Water System is serviced, unplug the Electrical Power Cord. The Water System must be powered OFF before plugging in or removing any components to the PCB(s). Other information Tools The following tools are referenced in this document and might be needed : • • • • Torx screwdrivers to disassemble and assemble systems and subassemblies. Diagnostic box EDI Resistor (150Ω) to be plugged on the EDI Board (load used to replace the EDI Module) UV Resistor (50Ω) to be plugged on the Main Control PCB (load used to replace the UV lamp 10W or 6W) Trademarks Millipore is a registered trademark of EMD Millipore Corporation. Millitrack software and the “M” logo are trademarks of EMD Millipore Corporation. Internet Explorer and Windows are registered trademarks of Microsoft Corporation, Adobe and Flash are registered trademarks of Adobe Systems Incorporated. All other trademarks are trademarks of their respective manufacturers. Copyright © 2012 EMD MILLIPORE CORPORATION. PRINTED IN FRANCE. ALL RIGHTS RESERVED. THIS BOOK OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHERS Document V2.0 – 07/12 Date of creation July 2012 SM_ESSENTIAL RANGE_TC&EC TABLE OF CONTENTS SM_ESSENTIAL RANGE_TC&EC Document usage .............................................................................................................. 1 MMI Navigation .............................................................................................................. 3 Software Navigation Maps ............................................................................................. 6 Summary of Troubleshooting Cards .............................................................................. 8 TC001 – Er:03 = RO Permeate Water Conductivity > SP ........................................... 11 TC002 - RO Low Feed Water Pressure ......................................................................... 13 TC003 - PROGARD/POLISHER PACK OUT OR INCORRECT ........................................ 15 TC004 – Er:30 = Coldfire is not responding ................................................................ 17 TC005 – Er:23 = PRESSURE SENSOR FAILURE (Distribution) ................................... 19 TC006 – Elix Product Resistivity < SP ........................................................................ 21 TC006 – Elix Product Resistivity < SP, Continued ..................................................... 22 TC007 - Dist Res < SP .................................................................................................. 23 TC008 – Er:02 = Feed Temperature > SP .................................................................... 25 TC009 – Er:05 = Feed Cond < MIN .............................................................................. 27 TC010 – Er:06 = Feed Cond > MAX ............................................................................ 29 TC011 – Er:07 = Permeate Temp < MIN ...................................................................... 31 TC012 – Er:08 = Permeate Temp > MAX .................................................................... 33 TC013 – Er:09 = Permeate Cond < MIN ...................................................................... 35 TC014 – Er:10 = Permeate Cond > MAX ..................................................................... 37 TC015 – Er:11 = Elix Temp < MIN .............................................................................. 39 TC016 – Er:12 = Elix Temp > MAX ............................................................................. 41 TC017 – Er:13 = Elix Res < MIN .................................................................................. 43 TC018 – Er:14 = Elix Res > MAX ................................................................................ 45 TC019 – Er:15 = Distribution Temp < MIN ................................................................. 47 TC020 – Er:16 = Distribution Temp > MAX ................................................................ 49 TC021 – Er:17 = Distribution Res < MIN .................................................................... 51 TC022 – Er:18 = Distribution Res > MAX ................................................................... 53 TC023 – Er:25 = Check UV Lamp ................................................................................ 55 TC024 – Er:27 = Check ASM UV Lamp ....................................................................... 57 TC025 – Er:01 = Tap Feed Cond > SP ......................................................................... 59 TC026 – Er:04 = RO rejection < SP ............................................................................. 61 Summary of Electronic Cards ....................................................................................... 64 General explanation - EC000 ....................................................................................... 65 Diagnostic Box – EC001 ............................................................................................... 66 Main Control PC Board – EC002 .................................................................................. 71 SM_ESSENTIAL RANGE_TC&EC Feed Conductivity Cell - EC003 ................................................................................... 75 RO Permeate Thermistor – EC004 ................................................................................ 79 RO Permeate Conductivity Cell – EC005 ..................................................................... 84 EDI Resistivity Cell – EC006 ........................................................................................ 89 Distribution Resistivity Cell – EC007 .......................................................................... 94 RO Pressure Sensor – EC008 ...................................................................................... 101 Distribution Pressure sensor – EC009 ........................................................................ 105 RO Inlet Solenoid Valve – EC010 .............................................................................. 108 RO Flush Solenoid Valve – EC011 ............................................................................. 114 RO Permeate Divert Solenoid Valve – EC012 ............................................................ 120 Recirculation Solenoid Valve – EC013 ...................................................................... 125 254nm UV Lamp – EC014 .......................................................................................... 130 Distribution Pump High Flowrate – EC015 ............................................................... 138 Distribution Pump Low Flowrate – EC016 ................................................................ 143 RO Pump – EC017 ....................................................................................................... 147 MAIN DISPLAY – EC018 ............................................................................................ 153 RFID Reader – EC019 ................................................................................................. 156 EDI Module – EC020 ................................................................................................... 159 Harness – EC021 ......................................................................................................... 165 External Power Supply – EC022 ................................................................................ 169 External Relay – EC023 .............................................................................................. 170 SM_ESSENTIAL RANGE_TC&EC Document usage Introduction Definition This Service Manual contains Troubleshooting and Electronic Components information for the ELIX/RIOS/AFS-D/AFS-E systems – Essential Range. Troubleshooting Card • A Troubleshooting Card (TC) is represented as a diagram. This diagram describes a sequence of actions to follow. This aids in troubleshooting a particular problem with the water system. • Each action to complete is written in an Action Box (square box). Detailed information can be found by clicking on . Electrical component testing information can be found by clicking on . • Troubleshooting Cards are classified into categories depending on the nature of the issue (see the Summary of Cards table for more information). Electronic Components Card An Electronic Components Card (EC) describes an electrical component found in the water system. Each EC is divided into several topics including: • Description and location, • Structure and characteristics, • Electrical connections and software interaction, and • Circuit and component tests. Using TC When using a Troubleshooting Card, click on the icon link to go directly to detailed information. To return back to TC, use the Previous View button or the Bookmarks link of your Adobe Reader Continued on next page SM_ESSENTIAL RANGE_TC&EC 1-172 Document usage, Continued Water systems The water systems are composed of two distinctive parts : The first portion of the system is called the RO portion or the RO/EDI portion composition if the system includes an EDI Module. All the systems of the Essential range incorporate a RO or RO/EDI portion. The second portion is called the Distribution portion for systems that include a Distribution part (AFS). The Rios and Elix systems do not incorporate Distribution portion. RO/EDI Portion Distribution Portion SM_ESSENTIAL RANGE_TC&EC 2-172 MMI Navigation The information given below describes how to access to the system information and how to perform certain operations by using the MMI (Man Machine Interface). Pressing right arrow button from Standby to access maintenance and lifetime Etc… Left arrow button, turns in reverse. Pressing on while in CL2 gives you access to set points and configuration menu. Press on to enter SP:01. TANK REFILL Press on to enter SP:02. % REJECTION Press on to enter SP:03. Product Water RESISTIVITY Press on change with arrows and press on to enter dAtE. change with arrows and press on change with arrows and press on change with arrows and press on change with arrows and press on change with arrows and press on change with arrows and press on change with arrows and press on Press on to exit IP address A, b, C and d values. change with arrows and press on to exit page see next Continued on next page SM_ESSENTIAL RANGE_TC&EC 3-172 MMI Navigation, Continued Press on to enter SP:01. TANK REFILL press on enter to enter the code each digit with arrows and Pressing on while in configuration puts the system in Ready. Action Code 159 à Entering FSE parameter Menu tank choose tank volume with arrows and on tank empty level calibration low level volume level calbration tank full level 0: no filter final filter 1: Millipak 2: Biopak type 3: Opticap Action Code 369 à Starting RO Rinse sequence change with arrows and press on press on high the RO Rinse is launched upon code validation to start press on wait… until you have reached 1 hour and the cancel button appears press on for a few seconds to cancel and exit the RO Rinse back to ready Action Code 147 à Pretreatment Override (24h) and Action Code 258 à Polisher Override (24h) in Ready… code 147 in Standby… code 147 Lab closed activated press on to go in Standby Action Code 357 à Reset of the Distribution pump lifetime Continued on next page SM_ESSENTIAL RANGE_TC&EC 4-172 MMI Navigation, Continued Operation parameters – in Ready or Tank Filling Tap Feed Conductivity (Elix, RiOs, AFS HF, AFS E) RO Feed Conductivity (Elix, RiOs, AFS HF, AFS E) RO Temperature (Elix, RiOs, AFS HF, AFS E) RO Pressure (Elix, RiOs, AFS HF, AFS E) % RO Rejection (Elix, RiOs, AFS HF, AFS E) RO Permeate Conductivity (Elix, AFS HF, AFS E) EDI Product Resistivity or conductivity (AFS E) EDI Product Temperature (AFS E) Lifetimes reset on press to display value to exit press on final filter (AFS) LF:02: UV lamp LF:03: ASM UV lamp (ElixRiOs) Alerts press on to access CL2 cleaning Alarms press on to start the cleaning press on to cancel the error code and go back to Ready. pack out or incorrect pack is blinking and system is stopped + water detected. Press on . Erreur 23 SM_ESSENTIAL RANGE_TC&EC 5-172 Software Navigation Maps RIOS/ELIX/AFS May be different depending on the system type and the accessories availability Continued on next page SM_ESSENTIAL RANGE_TC&EC 6-172 TROUBLESHOOTING CARDS SM_ESSENTIAL RANGE_TC&EC 7-172 Summary of Troubleshooting Cards Card MMI Error codes or icons TC001 Title Value RO Permeate Water Conductivity > SP Default setpoint = 100µS/cm (with EDI) 200 μS/cm (without EDI) <0.5 bar in Pressurization mode Low Feed Water Pressure TC002 TC003 Progard/ Polisher Pack is out or incorrect TC004 Coldfire is not responding TC005 Distribution Pressure sensor failure TC006 Elix product res < SP Default setpoint = 1MΩ.cm@25°C Dist res < SP Default setpoint = 10 MΩ.cm@25°C TC007 SM_ESSENTIAL RANGE_TC&EC • • • RO performance . Setpoint value. Calibration. <3.5 bar during 1 min in Tank Filling mode for systems with EDI <2.5 bar during 1 min in Tank Filling mode for systems without EDI or Description Progard or Polisher Pack TAG not detected by TAG Reader PCB. The Coldfire microprocessor is not responding when interrogated Pressure sensor not powered in recirculation or distribution modes. • Cell issue • Closed circuit • Elix water resistivity setpoint value • Cell issue. • Closed circuit • Replace Polisher cartridge. • Setpoint value. • Calibration 8-172 Card MMI Error codes or icons Title Value TC008 Feed Temp > SP Default setpoint value = 35°C TC009 Feed cond < MIN Min value : 4 μS/cm TC010 Feed cond > MAX Max value : 5000 μS/cm TC011 Permeate temp < MIN RO Permeate temp < 0°C TC012 Permeate temp > MAX RO Permeate temp > 50°C TC013 Permeate cond < MIN TC014 Permeate cond > MAX TC015 Elix temp < MIN Elix temp < 0°C TC016 Elix temp > MAX Elix temp > 50°C TC017 Elix res < MIN MIN Value : 0.17 MΩ.cm@25°C TC018 Elix res > MAX MAX Value : 200 MΩ.cm@25°C TC019 Distribution temp < MIN Distrib temp < 0°C SM_ESSENTIAL RANGE_TC&EC Permeate Conductivity < 0.26 µS/cm Permeate Conductivity > 250 µS/cm Description • Thermistor issue • Closed circuit • Calibration • Cell issue • Open circuit • Calibration • Cell issue • Closed circuit on FEED • Feedwater conductivity • Calibration • Thermistor issue • Open circuit • Calibration • Thermistor issue • Closed circuit • Permeate water temperature • Calibration • Cell issue. • Open circuit • Calibration • Cell issue. • Closed circuit • Calibration • Thermistor issue • Open circuit • Calibration. • Thermistor issue. • Closed circuit • Elix water temperature • Calibration. • Cell issue. • Closed circuit • Calibration • Cell issue. • Open circuit • Calibration • Thermistor issue • Open circuit • Calibration 9-172 Card MMI Error codes or icons Title Value TC020 Distribution temp > MAX Distrib temp > 50°C TC021 Distribution res < MIN MIN Value : 0.17 MΩ.cm@25°C TC022 Distribution res > MAX MAX Value : 200 MΩ.cm@25°C TC023 Check UV Lamp TC024 Check ASM UV Lamp TC025 Tap Feed Cond > SP TC026 RO rejection < SP SM_ESSENTIAL RANGE_TC&EC Default setpoint value : 2000 µS/cm@25°C Default setpoint value : 92% Description • Thermistor issue. • Closed circuit • Elix water temperature • Calibration • Cell issue. • Closed circuit • Calibration • Cell issue. • Open circuit • Calibration Intensity from UV Ballast is < 147 mA. Intensity from UV Ballast is < 147 mA • Feedwater conductivity • Setpoint value Calibration • Feedwater conductivity • RO membrane performance 10-172 TC001 – Er:03 = RO Permeate Water Conductivity > SP Type of message : Alarm Stop Default Set Point value : 100 μS/cm @25°C (system with EDI) 200 μS/cm @25°C (system without EDI) Why? The RO Permeate Water Conductivity is > SP in either Rinsing mode or Tank Filling mode. The Red Alarm Stop is raised and the system goes to Alarm Stop sequence. When? It takes 5 continuous minutes of the condition above for this to occur. What happened? The RO or RO/EDI portion totally stops to operate. The independent Distribution portion (AFS) allows to deliver water until tank is empty. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 6 Action Connect system to Millitrack Verify set points and values in Millitrack Verify Feed water Conductivity. If the Feedwater Conductivity is really a very high value then it is possible for the Permeate Conductivity to be very high. Verify that a SYSTEM RINSE was done (the RO membrane is rinsed out completely). Verify that a sanitant or cleaning chemical (ROCLEAN-A,...) is not in the RO Housing. Check for a closed circuit on the permeate conductivity cell. Use a Voltmeter to check the impedance of the wires for a closed circuit. A particle might exist between the 2 electrodes of the Conductivity. Cell causing a very conductive circuit. Check the O-rings on the permeate stem of the RO membrane inside the RO Housing. Continued on next page SM_ESSENTIAL RANGE_TC&EC 11-172 TC001 – Er:03 = RO Permeate Water Conductivity > SP, Continued SM_ESSENTIAL RANGE_TC&EC 12-172 TC002 - RO Low Feed Water Pressure Type of message : Alarm Stop Why? This icon is displayed when pressure measured is : • • • <0.5 bar during 1 min in Pressurization mode, or <3.5 bar during 1 min in Tank Filling mode for systems with EDI, or <2.5 bar during 1 min in Tank Filling mode for systems without EDI When? It takes 1 minute of the condition above for this to occur. What happened? The microprocessor has detected a signal back from the pressure sensor circuit which is less than the Pressure Setpoint. The measured pressure was < the Pressure Setpoint for > 2-3 minutes. The Pressure Setpoint is normally set to 0.5 Bar. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 Action Verify possible hydraulics causes : • Feed water off • External or internal tubings kinked • Progard clogged • Cleaning port clogged • Check RO Reject Flow Verify possible component and/or sensor causes : • RO Pump connected the wrong way • Int Pressure Regulator leakage or incorrect setting • Inlet strainer • Inlet Solenoid Valve closed • RO Reject and Permeate valves status and operation • Solenoid valves wiring • Reject and Permeate valves switched Continued on next page SM_ESSENTIAL RANGE_TC&EC 13-172 TC002 - RO Low Feed Water Pressure, Continued SM_ESSENTIAL RANGE_TC&EC 14-172 TC003 - PROGARD/POLISHER PACK OUT OR INCORRECT Type of message : Alarm Stop Why? Progard/Polisher Tag not detected by the Tag Reader or wrong pack detected. The Red Alarm Stop or is displayed. When? It takes 5 seconds of the condition above for this to occur. What happened? The water system totally stops to operate. What to do? Follow the Troubleshooting Card hereafter to solve this dysfunction. Continued on next page SM_ESSENTIAL RANGE_TC&EC 15-172 TC003 - PROGARD/POLISHER PACK OUT OR INCORRECT, Continued PACK Visual Inspection Tag + sticker + catalog number : Verification OK? No TAG DIAGNOSTIC TOOL Change PACK Reading ok? No TAG Reader Version OK? (Millitrack àAdvanced parameters) No Change Main PCB Version OK? Yes Change PACK Yes Verify quality of contact between tag reader and Main Control PCB Replace cable + Reader with external reader + cable (bypass) No Yes Yes No Finalize replacement of cable + tag reader Repair connector or contact Contact OK? Yes Change Reader Note : To temporarily overcome this issue, a 24h override can be performed using Millitrack (Configuration tab) or the MMI Action Codes (C:147 for Pretreatment override/C:258 for Polisher override). SM_ESSENTIAL RANGE_TC&EC 16-172 TC004 – Er:30 = Coldfire is not responding Type of message : Alarm Stop If an error code Er:30 (Coldfire is not responding) is displayed and before implementing the below procedure, save the files data_persistence.bin, nor_sdf.bin or nand_sdf.bin if available. For in-depth troubleshooting analysis, those files will be necessary. Why? The Coldfire microprocessor is not responding when interrogated. This error message can occur especially during upload. When? It takes 3 continuous minutes of the condition above for this to occur. What happened? The water system totally stops to operate. What to do? Follow the Troubleshooting Card hereafter to solve this dysfunction. Continued on next page SM_ESSENTIAL RANGE_TC&EC 17-172 TC004 – Er:30 = Coldfire is not responding, Continued If a new config.ini from the software package is selected, then “PC” will be displayed on the screen. - SM_ESSENTIAL RANGE_TC&EC Cancel “PC” by pressing 3 seconds on Enter system calibration factors found in the test bench manager database 18-172 TC005 – Er:23 = PRESSURE SENSOR FAILURE (Distribution) Type of message : Alarm Stop Why? The pressure sensor located in the distribution portion is not powered anymore in recirculation or distribution modes. When? It takes 5 continuous minutes of the condition above for this to occur. What happened? The water system and the distribution process stop to operate. What to do? Follow the Troubleshooting Card hereafter to solve this dysfunction. Continued on next page SM_ESSENTIAL RANGE_TC&EC 19-172 TC005 – Er:23 = PRESSURE SENSOR FAILURE (Distribution), Continued SM_ESSENTIAL RANGE_TC&EC 20-172 TC006 – Elix Product Resistivity < SP àAFS-E/ à Elix Type of message : Alarm Default setpoint value : 1MΩ.cm@25°C Why? The Elix Product Resistivity is < SP in Tank Filling mode. The Alarm or is displayed. When? It takes 1 minute of the condition above for this to occur. What happened? When this alarm occurs, the system remains in Tank Filling mode. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 Action Verify setpoint setting and measured value in Millitrack Verify the proper operation of the EDI Module. This includes: checking that the EDI Module is getting a voltage and intensity • checking that the EDI waste stream is connected on the back of the system • checking that the RO cartridge is working properly (good % rejection) Check for a closed circuit on the Elix resistivity cell. Use a Voltmeter to check the impedance of the wires for a closed circuit (see EC006/Test of the Elix Resistivity cell). particles might exist between the 2 electrodes of the Conductivity cell causing a very conductive circuit. • 3 4 5 Try increasing (if possible) the intensity to the EDI Module by changing the setting. Use the Flow Factor to lower the flowrate through the EDI Module. This may allow more residence time for ions to be removed. Continued on next page SM_ESSENTIAL RANGE_TC&EC 21-172 TC006 – Elix Product Resistivity < SP, Continued Verify setpoint setting and measured value in Millitrack Dispense water for several minutes Confirm resistivity and temperature measurement value with external meter No Measurement valid? Check cell calibration in Millitrack Yes - Verify Product flow - Verify Waste flow - Measure CO2 levels - Verify RO performance - Put the system in Production mode for several hours Check EDI PS Board setting No Correct setting? Change setting and let the system run for several minutes Yes Perform test with diagnostic box No Test ok? Yes Perform test with a resistor in place of the sensor Replace Main PCB No Replace harness SM_ESSENTIAL RANGE_TC&EC Test ok? Yes Replace resistivity cell 22-172 TC007 - Dist Res < SP Type of message : Alarm Default set point value : 10 MΩ.cm@25°C Why? The Distribution Resistivity is measured in AFSD/AFSD UV/AFS HF/AFS E HF water systems. The Distribution Resistivity is < SP in Recirculation or Dispensing modes. The Alarm Icon is displayed. When? It takes 10 seconds of the condition above for this to occur. What happened? The water system continues to operate. The water can still be dispensed. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 Action Connect system to Millitrack Verify set point and values in Millitrack Check for a closed circuit on the distribution resistivity cell. Use a voltmeter to check the impedance of the wires for a closed circuit (see EC007/Test of the Distribution Resistivity Cell). Particles might exist between the 2 electrodes of the Conductivity cell causing a very conductive circuit. Perform test with the diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 23-172 TC007 - Dist Res < SP, Continued Verify set point setting and measured value in Millitrack Confirm resistivity measurement value with external meter No No Measurement valid? Yes Check cell calibration in Millitrack Replace Polisher Perform test with Diagnostic box Calculate expected water throughput, compare to actual throughput Test ok? Yes Perform test with a resistor in place of the sensor Replace MAIN PCB No Replace harness SM_ESSENTIAL RANGE_TC&EC Test ok? Yes Replace resistivity cell 24-172 TC008 – Er:02 = Feed Temperature > SP Type of message : Alarm Default setpoint value : 35°C Why? The Feed water Temperature is > SP. When? The ERROR 2 message appears after 1 minute of RO : Flush, Rinsing or Tank Filling Modes. What happened? The measured Feed water temperature was > 35 °C for > 1 minute. The water system continues to operate. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 Action Check temperature with calibrated thermometer Verify the continuity of the thermistor wires (no short circuit) Check the impedance of the thermistor by itself. A closed circuit (0 Ω) will cause a display of > 45 °C. Perform test with the diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 25-172 TC008 – Er:02 = Feed Temperature > SP, Continued SM_ESSENTIAL RANGE_TC&EC 26-172 TC009 – Er:05 = Feed Cond < MIN Type of message : alarm Min value : 4 μS/cm Why? The Feed water conductivity is not within the expected range. When? The ERROR 5 message appeared after at least 10 seconds of RO : Flush, Rinsing or TankFilling Modes. The water system continues to operate. What happened? The measured Feedwater Conductivity was < 4 μS/cm for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 Action Check if the feedwater to the water system is really < 4 µS/cm NTC (i.e. distilled feedwater, DI feedwater). Perform test with the diagnostic box. Make sure the displayed feedwater conductivity is 0 µS/cm for an open circuit and > 3389 µS/cm for a closed circuit on Feed cond cell Check for an open circuit on the feedwater conductivity cell (the wires plug into the RO Housing Head). Use a Voltmeter to check the impedance of the wires for an open circuit (see EC003 – Test of the Feed Cond Cell). Continued on next page SM_ESSENTIAL RANGE_TC&EC 27-172 TC009 – Er:05 = Feed Cond < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 28-172 TC010 – Er:06 = Feed Cond > MAX Type of message : alarm Max value : 4420 μS/cm Why? The Feed water conductivity is not within the expected range. When? The ERROR 6 message appeared after at least 10 seconds of RO : Flush, Rinsing or Tank Filling Modes. The water system continues to operate. What happened? The measured Feedwater Conductivity was > 4420 µS/cm for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 Action Check if the feedwater to the water system is really > 4420 µS/cm NTC. 2 Perform test with the diagnostic box. Make sure the displayed feedwater conductivity is 0 µS/cm for an open circuit and > 3389 µS/cm for a closed circuit on Feed cond cell Check for a closed circuit on the feedwater conductivity cell (the wires plug into the RO Housing Head). Use a Voltmeter to check the impedance of the wires for a closed circuit (see EC003 – Test of the Feed Cond Cell). 3 Continued on next page SM_ESSENTIAL RANGE_TC&EC 29-172 TC010 – Er:06 = Feed Cond > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 30-172 TC011 – Er:07 = Permeate Temp < MIN Type of message : alarm Min value : 0°C Why? The Permeate water temperature is not within the expected range. When? The ERROR 7 message appeared after at least 10 seconds of RO Rinsing or Tank Filling Modes. The water system continues to operate. What happened? The Permeate water Temperature was < 0°C for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 Action Check the impedance of thermistor by itself. An open circuit (∞ Ω) will cause a display of < 0 °C. Perform test with diagnostic box Perform test with a resistor in place of the thermistor. A 100 KΩ resistor should yield a displayed temperature of 25 °C Continued on next page SM_ESSENTIAL RANGE_TC&EC 31-172 TC011 – Er:07 = Permeate Temp < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 32-172 TC012 – Er:08 = Permeate Temp > MAX Type of message : alarm Max value : 50°C Why? The Permeate water temperature is not within the expected range. When? The ERROR 8 message appeared after at least 10 seconds of RO : Rinsing or Tank Filling Modes. What happened? The Permeate water temperature was > 50°C for > 10 seconds. The water system continues to operate. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Confirm temperature value with external meter Check the thermistor wires Check the impedance of the thermistor by itself. A closed circuit (0 Ω) will cause a display of > 45 °C. Perform test with diagnostic box Perform test with a resistor in place of the thermistor. A 100 KΩ resistor should yield a displayed temperature of 25 °C Continued on next page SM_ESSENTIAL RANGE_TC&EC 33-172 TC012 – Er:08 = Permeate Temp > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 34-172 TC013 – Er:09 = Permeate Cond < MIN Type of message : Alarm Min value : 0.26 µS/cm Why? The Permeate water conductivity is not within the expected range. When? The ERROR 9 message appeared after at least 10 seconds of RO : Rinsing or Tank Filling Modes. The water system continues to operate. What happened? The measured Permeate Conductivity was < 0.26 µS/cm for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step Action Check if the feedwater conductivity to the water system is extremely low (i.e. DI feedwater). 2 Check the connection of the Permeate Conductivity Cell Check for an open circuit on the permeate conductivity cell. Use a Voltmeter to check the impedance of the wires for an open circuit (see EC005 – Test of the Permeate conductivity cell) 1 3 4 Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 35-172 TC013 – Er:09 = Permeate Cond < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 36-172 TC014 – Er:10 = Permeate Cond > MAX Type of message : Alarm Max value : 310 µS/cm Why? The Permeate water conductivity is not within the expected range. When? The ERROR 10 message appeared after at least 10 seconds of RO : Rinsing or Tank Filling Modes. The water system continues to operate. What happened? The measured Permeate Conductivity was > 310 µS/cm for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 Action Verify that a system Rinse was done (the RO membrane is rinsed out completely). Verify that a sanitant or cleaning chemical (RO Clean A,…) is not in the RO Housing. Check the connection of the Permeate Conductivity Cell 3 Check for a closed circuit on the permeate conductivity cell. Use a Voltmeter to check the impedance of the wires (see EC005 – Test of the Permeate conductivity cell). A particle might exist between the electrodes of the conductivity cell causing a very conductive circuit. 4 Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 37-172 TC014 – Er:10 = Permeate Cond > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 38-172 TC015 – Er:11 = Elix Temp < MIN Type of message : Alarm MIN Value : 0°C Why? The Elix water temperature is not within the expected range. When? The ERROR 11 message appeared after at least 10 seconds of Tank Filling Mode. The water system continues to operate. What happened? The Permeate water temperature was < 0°C for > 10 seconds. The water system continues to operate. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Confirm temperature value with external meter Check the thermistor wires Check the impedance of the thermistor by itself. An open circuit (∞ Ω) will cause a display of < 0 °C. Perform test with a resistor in place of the thermistor. A 100 KΩ resistor should yield a displayed temperature of 25 °C Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 39-172 TC015 – Er:11 = Elix Temp < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 40-172 TC016 – Er:12 = Elix Temp > MAX Type of message : Alarm Max Value : 50°C Why? The Elix water temperature is not within the expected range. When? The ERROR 12 message appeared after at least 10 seconds of Tank Filling Mode. The water system continues to operate. What happened? The Elix water temperature was > 50°C for > 10 seconds. The water system continues to operate. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Confirm temperature value with external meter Check the thermistor wires Check the impedance of the thermistor by itself. A closed circuit (0 Ω) will cause a display of > 45 °C. Perform test with a resistor in place of the thermistor. A 100 KΩ resistor should yield a displayed temperature of 25 °C Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 41-172 TC016 – Er:12 = Elix Temp > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 42-172 TC017 – Er:13 = Elix Res < MIN Type of message : Alarm MIN Value : 0.17 MΩ.cm@25°C Why? The Elix water resistivity is not within the expected range. When? The ERROR 13 message appeared after at least 10 seconds of Tank Filling Mode. The water system continues to operate. What happened? The measured Elix Resistivity was < 0.17 MΩ.cm@25°C for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Connect system to Millitrack Verify value in Millitrack Check the connection of the Elix Resistivity Cell Check for a closed circuit on the Elix Resistivity cell. Use a Voltmeter to check the impedance of the wires (see EC005 – Test of the Permeate conductivity cell). A particle might exist between the electrodes of the conductivity cell causing a very conductive circuit. Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 43-172 TC017 – Er:13 = Elix Res < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 44-172 TC018 – Er:14 = Elix Res > MAX Type of message : Alarm MAX Value : 200 MΩ.cm@25°C Why? The Elix water resistivity is not within the expected range. When? The ERROR 14 message appeared after at least 10 seconds of Tank Filling Mode. The water system continues to operate. What happened? The measured Elix Resistivity was > 200 MΩ.cm@25°C for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 Action Connect system to Millitrack Verify value in Millitrack Check the connection of the Elix Resistivity Cell 4 Check for an open circuit on the Elix Resistivity cell. Use a Voltmeter to check the impedance of the wires for an open circuit (EC006 – Test of the EDI Resistivity cell 5 Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 45-172 TC018 – Er:14 = Elix Res > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 46-172 TC019 – Er:15 = Distribution Temp < MIN Type of message : Alarm MIN Value : 0°C Why? The Product water temperature is not within the expected range. When? The ERROR 15 message appeared after at least 10 seconds of Distribution Mode. The water system continues to operate. What happened? The Elix water temperature was < 0°C for > 10 seconds. The water system continues to operate. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Confirm temperature value with external meter Check the thermistor wires Check the impedance of the thermistor by itself. An open circuit (∞ Ω) will cause a display of < 0 °C. Perform test with a resistor in place of the thermistor. A 100 KΩ resistor should yield a displayed temperature of 25 °C Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 47-172 TC019 – Er:15 = Distribution Temp < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 48-172 TC020 – Er:16 = Distribution Temp > MAX Type of message : Alarm MAX Value : 50°C Why? The Product water temperature is not within the expected range. When? The ERROR 16 message appeared after at least 10 seconds of Distribution Mode. The water system continues to operate. What happened? The Distribution water temperature was > 50°C for > 10 seconds. The water system continues to operate. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Confirm temp value with external meter Check the thermistor wires Check the impedance of the thermistor by itself. A closed circuit (0 Ω) will cause a display of > 45 °C. Perform test with a resistor in place of the thermistor. A 100 KΩ resistor should yield a displayed temperature of 25 °C Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 49-172 TC020 – Er:16 = Distribution Temp > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 50-172 TC021 – Er:17 = Distribution Res < MIN Type of message : Alarm MIN Value : 0.17 MΩ.cm@25°C Why? The Product water resistivity is not within the expected range. When? The ERROR 17 message appeared after at least 10 seconds of Distribution Mode. The water system continues to operate. What happened? The measured Distribution Resistivity was < 0.17 MΩ.cm@25°C for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 5 Action Connect system to Millitrack Verify value in Millitrack Check the connection of the Distribution Resistivity Cell Check for a closed circuit on the Elix Resistivity cell. Use a Voltmeter to check the impedance of the wires (see EC007 – Test of the Distribution Resistivity cell). Particles might be present between the electrodes of the conductivity cell causing a very conductive circuit. Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 51-172 TC021 – Er:17 = Distribution Res < MIN, Continued SM_ESSENTIAL RANGE_TC&EC 52-172 TC022 – Er:18 = Distribution Res > MAX Type of message : Alarm MAX Value : 200 MΩ.cm@25°C Why? The Product water resistivity is not within the expected range. When? The ERROR 18 message appeared after at least 10 seconds of Distribution Mode. The water system continues to operate. What happened? The measured Distribution Resistivity was > 200 MΩ.cm@25°C for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 Action Connect system to Millitrack Verify value in Millitrack Check the connection of the Distribution Resistivity Cell 4 Check for an open circuit on the Distribution Resistivity cell. Use a Voltmeter to check the impedance of the wires for an open circuit (EC007 – Test of the Distribution Resistivity cell) 5 Perform test with diagnostic box Continued on next page SM_ESSENTIAL RANGE_TC&EC 53-172 TC022 – Er:18 = Distribution Res > MAX, Continued SM_ESSENTIAL RANGE_TC&EC 54-172 TC023 – Er:25 = Check UV Lamp Type of message : Alarm Message displayed when intensity value is below 147 mA Why? The UV Lamp intensity is not within the expected range. When? The ERROR 25 message appeared after at least 10 seconds of Distribution or Recirculation Modes. The water system continues to operate. What happened? The measured intensity to the UV lamp was < 147 mA for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 Action Verify Lamp UV, UV Ballast board and Main Board connection 2 Perform test with UV Resistor (50Ω) connected to the J7 Main board connector (See EC001 – Tools needed to perform system test) Test with light bulb in place of the UV Lamp 3 Continued on next page SM_ESSENTIAL RANGE_TC&EC 55-172 TC023 – Er:25 = Check UV Lamp, Continued SM_ESSENTIAL RANGE_TC&EC 56-172 TC024 – Er:27 = Check ASM UV Lamp Type of message : Alarm Message displayed when intensity is below 147 mA Why? The UV Lamp intensity is not within the expected range. When? The ERROR 27 message appeared after at least 10 seconds of operation. The water system continues to operate. What happened? The measured intensity to the UV lamp was < 147 mA for > 10 seconds. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 Action Verify lamp UV, UV Ballast board and Main Board connection 2 Perform test with UV Resistor (50Ω) connected to the connector J8 of the Main board (See EC001 – Tools needed to perform system test) Test with light bulb in place of the UV Lamp 3 Continued on next page SM_ESSENTIAL RANGE_TC&EC 57-172 TC024 – Er:27 = Check ASM UV Lamp, Continued SM_ESSENTIAL RANGE_TC&EC 58-172 TC025 – Er:01 = Tap Feed Cond > SP Type of message : Alert Default setpoint value : 2000 µS/cm@25°C Why? The Tap Feed Conductivity is > SP. When? The Tap Feed Conductivity measurement is performed when the system is in Flush mode. The ERROR 1 message appeared after at least 1 minute of Flush Mode. The water system continues to operate. What happened? The Tap Feed Conductivity was > SP for > 1 minute. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 2 3 4 Action Connect system to Millitrack Verify set points and values in Millitrack Verify the Tap Feed Conductivity with external meter Verify the Feed Conductivity Cell (see EC003 – Test of the Feed Cond Cell) 5 To reset the Er:01 alert, first make sure that the Feed Conductivity is < SP. Then, the reset will be done automatically the next Flush (every 5 hours) or power Off/On the system. Continued on next page SM_ESSENTIAL RANGE_TC&EC 59-172 TC025 – Er:01 = Tap Feed Cond > SP, Continued SM_ESSENTIAL RANGE_TC&EC 60-172 TC026 – Er:04 = RO rejection < SP Type of message : Alert Default setpoint value : 92% Why? The RO Rejection % is < SP. When? The ERROR 4 message appeared after at least 5 continuous minutes of Rinsing or Tank Filling Mode. The water system continues to operate. What happened? The measured % Rejection was < the % Rejection Setpoint for > 5 continuous minutes. What to do? Follow the action steps below and the Troubleshooting Card to solve this dysfunction : Step 1 Action Verify the % Rejection in Millitrack 2 3 Operate the system for 15-20 minutes Verify the Feed Conductivity (very low feedwater conductivity can lead to low % Rejection), the Permeate Conductivity, the pump pressure, the RO Permeate flowrate, the RO Reject flowrate 4 Check the presence of kinks or obstructions to the RO Reject flow 5 Check if the Progard has been changed recently. a new Progard will release minerals from the activated carbon and/or the polyphosphate. This can lead to a lowering of the % rejection. 6 Perform an ACID (ROCLEAN-A) and/or a BASE Cleaning (ROCLEAN-B) 7 Verify that the O-rings on the RO membrane are well seated. Continued on next page SM_ESSENTIAL RANGE_TC&EC 61-172 TC026 – Er:04 = RO rejection < SP, Continued SM_ESSENTIAL RANGE_TC&EC 62-172 ELECTRONIC CARDS SM_ESSENTIAL RANGE_TC&EC 63-172 Summary of Electronic Cards ELECTRONIC COMPONENTS Card# EC000 EC001 EC002 EC003 EC004 EC005 EC006 EC007 EC008 EC009 EC010 EC011 EC012 EC013 EC014 EC015 EC016 EC017 EC018 EC019 EC020 EC021 EC022 EC023 Type General explanation Diagnostic Box Main Control PC Board Feed Conductivity Cell RO Permeate Thermistor RO Permeate Conductivity Cell EDI Resistivity Cell Distribution Resistivity Cell RO Pressure Sensor Distribution Pressure sensor RO Inlet Solenoid Valve RO Flush Solenoid Valve RO Permeate Divert Solenoid Valve Recirculation Solenoid Valve 254 nm UV LAMP High Flowrate Distribution Pump Low Flowrate Distribution Pump RO Pump Main Display RFID Reader EDI Module Harness Assembly External power supply External Relay Content • Description • Location • Part number • Structure • Characteristics • Electrical connections • Software interaction • Circuit test • Component test Legend: MMI: Man Machine Interface MC PCB: Main Control Printed Circuit Board PCB: Printed Circuit Board UV: Ultra Violet SM_ESSENTIAL RANGE_TC&EC 64-172 General explanation - EC000 Introduction The section below describes how to do 2 tests that appear in almost all troubleshooting cards: • Test of electrical continuity, • Verification of contact quality. Test of continuity A continuity test checks if a circuit is electrically continuous thus eliminating the possibility of an open circuit. It is useful when you want to explore a circuit board, test connections, or make sure your soldering joints are all effective. In this document, a continuity test is used to eliminate the possibility of a break or open circuit in various cables and such. Proceed as follows: Verification of contact quality Step 1 2 Action Turn your multi-meter on. Turn it to the continuity setting which looks like a little speaker: 3 Test it by touching the probes together and you should hear a beep. That’s the same beep that you’ll hear if you test two connections that are connected! In most of the troubleshooting cards, the diagram starts with “Verify quality of contact”. Check the following things: • Wires are securely inserted in the connector, • The Molex connector pins make good contact with the PCB, • The Molex connector is located correctly. SM_ESSENTIAL RANGE_TC&EC 65-172 Diagnostic Box – EC001 Description The diagnostic box is used by the Field Service Engineer in order to investigate the system. The diagnostic box allows to simulate all the sensors measurements by means of rotary switches. It is also used to perform the system calibration with NIST resistors. The diagnostic box allows to simulate different signals according to the rotary switches position : • • • • • • • • • • The Feed Conductivity The Permeate Conductivity The Elix Resistivity The Distribution Resistivity The Feed/Permeate Temperature The Elix Temperature The Distribution Temperature The RO Pressure The Distribution Pressure The tank level The different switches positions are described below : • • • • • Two nominal values : N1&N2 Con : This position allows to simulate the sensor by plugging a specific resistor (NIST) directly on the diagnostic box (bottom right). It is used mainly to calibrate the system. Var : This position allows to simulate and modify the sensor signal by turning the corresponding rotary switch. OC : open circuit position CC : closed circuit position Continued on next page SM_ESSENTIAL RANGE_TC&EC 66-172 Diagnostic Box – EC001, Continued Description (continued) Electrical connection The different sensors values simulated with the diagnostic box are the following : : The system must be in Standby mode before connecting the diagnostic box. The diagnostic box is connected to the main PCB through the J12 connector. • Remove the sensors connector and plug the diagnostic box connector instead • Plug the tank level connector to the system Tank level connection Continued on next page SM_ESSENTIAL RANGE_TC&EC 67-172 Diagnostic Box – EC001, Continued How to use the Put all the rotary switches in nominal value N1 position. diagnostic box To go in tank filling mode, the depressurization and pressurization modes must be simulated with the diagnostic box by modifying the RO pressure. Step 1 Action When the system goes in Depressurization, turn the RO Pressure rotary switch (Var position) located on the diagnostic box all the way to the left. Continued on next page SM_ESSENTIAL RANGE_TC&EC 68-172 Diagnostic Box – EC001, Continued How to use the diagnostic box (continued) Step 2 Action When the system goes in Pressurization, turn the RO Pressure rotary switch (Var position) located on the diagnostic box all the way to the right. For the distribution operation, the Distribution Pressure rotary switch allows to simulate the Distribution Stop (N1 : 2.1 bar pressure simulation) or the Distribution Start (N2 : 0.5 bar pressure simulation). SM_ESSENTIAL RANGE_TC&EC 69-172 Diagnostic Box – EC001, Continued Tools needed to perform system test - Diagnostic box - EDI resistor (150Ω) to be plugged on the EDI PS Board in place of the EDI Module (EDI Resistor will be powered in Tank Filling mode) - UV resistor (50Ω) to replace 10W or 6W UV lamp to be plugged on the connector J7 of the Main PC board (UV Resistor will be powered during Distribution mode). HEAT IS DISSIPATED BY THE RESISTORS/LOCATE THE RESITORS IN A SAFE AREA (AWAY FROM WATER) WHEN CONNECTED TO THE SYSTEM SM_ESSENTIAL RANGE_TC&EC 70-172 Main Control PC Board – EC002 Description The Main Control PC Board is in charge of several functions : - Location Control the process thanks to a µC Coldfire Control the actuators (pumps, solenoid valves, UV ballast, EDI power supply…) Perform all measurements Manage Ethernet communication Log history and defaults Manage the display and keypad The main Control PC Board is located on the top of the system under the top cover. Continued on next page SM_ESSENTIAL RANGE_TC&EC 71-172 Main Control PC Board – EC002, Continued Electrical connection Connectors Main PCB The electrical connection on the Main Control PC Board is described below : Connector J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 J13 J14 J15 J16 J17 Use Power entry 24V EV Inlet EV Flush EV Rinsing (3ways) EV Recir EDI Power supply (CIPS) UV Ballast Distribution pump brushed Aquatech (AFS HF) or brushless (AFS D) /ASM or relay (RIOS/ELIX) RO Pump Water sensor Tank level Harness Cable LCD Keypad Ethernet RFID Pretreatment RFID Polisher Comments Used in AFS D UV Sensors or Diagnostic Box Used in RIOS/ELIX/AFS Used in AFS Continued on next page SM_ESSENTIAL RANGE_TC&EC 72-172 Main Control PC Board – EC002, Continued Test of the Main Control PC Board To make sure that the Main Control PCB is fully operational, check the status of the LEDs in normal operation. LEDs 1 & 2 LED 5 LEDs 6/7/8 LED Description 1 2 3 4 5 µC PIC µC PIC Coldfire Coldfire µC PIC 6 7 8 24 V 3.3 V 5V Status in normal operation On Off Not used Not used On for 100 Mbits/s Ethernet On On On Continued on next page SM_ESSENTIAL RANGE_TC&EC 73-172 Main Control PC Board – EC002, Continued Voltage used by components The Voltage… 24V • • • • • 3.3V 5V • • • is used by… Solenoid valves Low & High flow pumps RO Pump EDI Board UV Ballast board Display Sensors Backlight RFID J12 Connector All the sensors and cells are connected to the Main PCB via the harness connected to the J12 connector. The pins identification is shown below : pins identification 9 19 9 20 1 2 10 J12 connector pins 1/2 3/4 5(measure)/6 (0V) 7 (5V)/8 (measure)/9(0V) 10/11 12/13 14/15 16/17 18 (5V) /19 (0V)/20 (measure) SM_ESSENTIAL RANGE_TC&EC Sensor Feed conductivity cell Permeate conductivity cell Permeate temperature sensor RO Pressure sensor Distribution resistivity cell Distribution temperature sensor EDI resistivity cell EDI temperature sensor Distribution Pressure sensor 74-172 Feed Conductivity Cell - EC003 Description The same cell (Feed Conductivity Cell) is used to measure : • The feed water conductivity : measurement made at the end of the startup flush (no water recovery because the RO Flush Solenoid Valve is open), • The RO feed water conductivity (measurement made during tank filling). Location The Feed Conductivity Cell is located on the top of the 1st RO Housing Assembly, as shown below : Elix Continued on next page SM_ESSENTIAL RANGE_TC&EC 75-172 Feed Conductivity Cell - EC003, Continued Catalog Number • The Feed Conductivity Cell does not have a catalog number because it is part of the RO Housing Assembly. • To change a Feed Conductivity Cell, you have to change the RO Housing Assembly. Characteristics The Feed Conductivity Cell has a cell constant of 0.285 cm-1. Software Interaction The table below describes software interactions of the cell : Description [RO Feed Conductivity] [TAP Feed Conductivity] Note : the [RO Feed Conductivity] can be higher than the [Tap Feed Conductivity] because a part of the RO reject is re-circulated to the RO feed. • Define the setpoint value; when the feed water conductivity is above this value, the system displays a major alert. • Default Value: 2000 µS/cm @ 25°C, • Value Range: 0.00 µS/cm – 5000 µS/cm (@ 25°C). Display In the Millitrack Advanced Parameters tab (Measures), the TAP Feed Conductivity and the RO Feed Conductivity are displayed. In the Millitrack Configuration tab, the setpoint value of the Tap Feed Conductivity can be adjusted. Continued on next page SM_ESSENTIAL RANGE_TC&EC 76-172 Feed Conductivity Cell - EC003, Continued Diagnostic A problem with the Feed Conductivity Cell can cause the following symptoms : Message Type What it means TAP FEED CONDUCTIVITY > SP RO FEED C > MAX RO FEED C < MIN Major Alert feedwater conductivity > Set Point. Delay Before display 60 s Alarm RO feedwater conductivity 10 s > 3500 µs/cm *. Alarm RO feedwater conductivity 10 s < 2.5 µs/cm *. * Value not dependent of the temperature or the compensation mode. The list above is not exhaustive. Note: The Feed Conductivity Cell cannot display conductivity below 2.5 µs/cm or above 3500 µs/cm, it will display “_ _ _”. Continued on next page SM_ESSENTIAL RANGE_TC&EC 77-172 Feed Conductivity Cell - EC003, Continued Test of the FEED Circuit The test must be carried out with the diagnostic box as indicated below : • • Test of the Conductivity Cell Step 1 2 3 4 Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the RO/Feed conductivity circuit by using the corresponding rotary switches on the diagnostic box. Action Visual inspection of the 2 electrodes. See Main Control PC Board EC002 to identify the Feed conductivity cell pins on harness assembly J12 and perform a continuity test. Test the Feed Conductivity Cell by measuring the resistance between the 2 wires on the connector while it is disconnected from the Main board. The measured resistance should be very high. The measured resistance will climb during the time the Voltmeter is used. Perform test with a resistor in place of the conductivity cell to test the conductivity cell wires and the signal acquisition by the Main board. Step 1 2 3 4 5 Action Power off the system. Disconnect the conductivity cell connector. Knowing that the Feed Conductivity cell constant is equal to 0.285 cm-1, a resistor of 400Ω is similar to a conductivity of approximately 700µS/cm. Connect the resistor in place of the conductivity cell. Power on the system and verify the measured value with Millitrack Continued on next page SM_ESSENTIAL RANGE_TC&EC 78-172 RO Permeate Thermistor – EC004 Description The RO Thermistor is placed on the permeate of the last RO cartridge. It measures the RO Permeate water temperature. The system uses this value to compensate the Permeate conductivity. Location The RO Thermistor is located on the RO Permeate Cell. RIOS Continued on next page SM_ESSENTIAL RANGE_TC&EC 79-172 RO Permeate Thermistor – EC004, Continued Location AFS Catalog Number The RO Thermistor does not have a catalog number since it is part of the RO Permeate Cell. Characteristics The RO Thermistor has a resistivity of 100 KΩ at 25°C. Continued on next page SM_ESSENTIAL RANGE_TC&EC 80-172 RO Permeate Thermistor – EC004, Continued Software Interaction The table below describes software interactions of the cell: Descript on RO Permeate Temperature - Display The permeate temperature is shown : in the Millitrack Advanced parameters tab (Measures). - Diagnostic Test of the temp RO circuit In the dashboard A problem with the RO Thermistor can cause the following symptoms : Message Type What it means Er:07 (Permeate temp < min) Er:08 (Permeate temp>max) Alarm RO Permeate temp<0°C Alarm RO Permeate temp>50°C Delay Before display 10 sec 10 sec The test must be carried out with the diagnostic box as described below : • • Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the RO/Feed temp circuit by using the corresponding rotary switch on the diagnostic box. Continued on next page SM_ESSENTIAL RANGE_TC&EC 81-172 RO Permeate Thermistor – EC004, Continued Test of the thermistor To test the thermistor by itself, follow the steps below : Step 1 2 3 Action Power off the system. Disconnect the thermistor. Use an Ohmmeter and measure the impedance of the thermistor Ω Continued on next page SM_ESSENTIAL RANGE_TC&EC 82-172 RO Permeate Thermistor - EC004, Continued Test of the thermistor (continued) Step 4 Action Check that the measured impedance is similar to the values shown in the table below. Due to several factors, it may not be exactly the same value.The table below gives the impedance of the thermistor for different water temperatures : Temperature (°C) 1 5 10 15 20 25 30 35 40 45 50 Resistivity (kΩ) 351 269 207 161 126 100 79 63 51 41 33 5 Note: The calibration of the thermistor circuit can change the display of the temperature. Analysis: 6 If the measured then the thermistor impedance… is… • matches the temperature OK. of the water faulty. • does not match the temperature of the water Return to the Troubleshooting Card. • To test the thermistor wires and the signal acquisition by the Main board, connect a 100 KΩ resistor in place of the sensor. A 100 KΩ resistor is equivalent to a temperature of 25°C. SM_ESSENTIAL RANGE_TC&EC 83-172 RO Permeate Conductivity Cell – EC005 Description The Permeate Conductivity Cell is used to measure the RO permeate conductivity. Location The Permeate Conductivity Cell is located after the RO cartridges. RIOS Continued on next page SM_ESSENTIAL RANGE_TC&EC 84-172 RO Permeate Conductivity Cell – EC005, Continued Location AFS Catalog Number The Permeate Conductivity Cell catalog number is ZF3000517 Continued on next page SM_ESSENTIAL RANGE_TC&EC 85-172 RO Permeate Conductivity Cell – EC005, Continued Structure The Conductivity Cell is made of different subparts : 1 2 5 3 6 4 Number Part Function 1 Conductivity cell electrodes 2 Thermistor wire & connector 3 Water outlet (port size: 8 mm) Built-in Strainer Measures the resistivity of the water after the RO membranes. Measures the temperature of the water after the RO membranes. Water goes out 4 5 6 Not shown Water inlet (port size: 6 mm) O-Ring Check valve Protects both the cell and the flowmeter down stream. Strainer pore size = 190 µm Water goes in Catalog Number ZF3000517 Prevent leaks Prevent water from the tank to go in the Elix module Characteristics The Permeate Conductivity Cell has a cell constant of 0.03 cm-1 Continued on next page SM_ESSENTIAL RANGE_TC&EC 86-172 RO Permeate Conductivity Cell – EC005, Continued Software Interaction The table below describes software interactions of the cell : Description RO Permeate Conductivity Display The permeate conductivity is shown in the Millitrack Advanced parameters tab (Measures). Define the setpoint value; when the permeate water conductivity is above this value, the system displays a major alert. • Default Value: 100 μS/cm @ 25°C (ELIX/AFS-E), 200 μS/cm @ 25°C (RIOS/AFS without EDI Module) • Value Range: 0.0 μS/cm – 250 μS/cm (@25°C). In the Millitrack Configuration tab, the setpoint value of the Permeate Conductivity can be adjusted. Diagnostic A problem with the RO Permeate Conductivity Cell can cause the following symptoms: Message Type What it means Delay before display Alarm Permeate 5 min Er:03 or Stop Conductivity> SP (PermeateCond>SP) Er:09 (Permeate Alarm Permeate 10 sec Cond<Min) Conductivity<0µS/cm Er:10 (Permeate Alarm Permeate 30 sec Cond>Max) Conductivity> 250 µS/cm Test of the RO Permeate Conductivity circuit The test must be carried out with the diagnostic box as indicated below : • • Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the RO Permeate conductivity circuit by using the corresponding rotary switch on the diagnostic box. Continued on next page SM_ESSENTIAL RANGE_TC&EC 87-172 RO Permeate Conductivity Cell – EC005, Continued Test of the Conductivity Cell Step 1 2 3 4 Action Visual inspection of the 2 electrodes See Main Control PC Board EC002 to identify the Permeate conductivity cell pins on harness assembly J12 and perform a continuity test. Test the Permeate Conductivity Cell by measuring the resistance between the 2 wires on the connector while disconnected from the Main board. The measured resistance should be very high. The measured resistance will climb during the time the Voltmeter is used. Perform test with a resistor in place of the permeate conductivity cell to test the cell wires and the signal acquisition by the Main board. Step 1 2 3 4 5 SM_ESSENTIAL RANGE_TC&EC Action Power off the system. Disconnect the conductivity cell connector. Knowing that the Permeate Conductivity cell constant is equal to 0.03 cm-1, a resistor of 3 KΩ is similar to a conductivity of 10µS/cm. Connect the resistor between the electrodes. Power on the system and verify the measured value with Millitrack. 88-172 EDI Resistivity Cell – EC006 Description The EDI Resistivity Cell is a sensor that measures the water resistivity after the Elix module. The resistivity cell includes a thermistor in order to measure the temperature of the water. This temperature is used to compensate the Elix resistivity. Resistivity Cell Thermistor Location The EDI Resistivity Cell is used in Elix/Elix UV/AFSE systems ELIX Continued on next page SM_ESSENTIAL RANGE_TC&EC 89-172 EDI Resistivity Cell – EC006, Continued Location AFS-E Catalog Number The EDI Resistivity Cell catalog number is ZF3000517 Characteristics The table below describes the characteristics of the EDI Resistivity Cell : Cell constant 0.03 cm-1 Thermistor 100 KΩ@25°C Port sizes 6/8 mm Continued on next page SM_ESSENTIAL RANGE_TC&EC 90-172 EDI Resistivity Cell – EC006, Continued Software Interface Diagnostic The table below describes software interactions of the cell : Description Elix Resistivity Display The Elix Resistivity value is shown in the Millitrack Advanced parameters tab (Measures). Define the setpoint value; when the Elix water resistivity is above this value, the system displays a major alert. • Default Value: 1 MΩ.cm @ 25°C (ELIX/AFS-E), • Value Range: 0 MΩ.cm – 15 MΩ.cm (@25°C). In the Millitrack Configuration tab, the setpoint value of the Elix resistivity cell can be adjusted. A problem with the EDI Resistivity Cell can cause the following symptoms : Message Type What it means Delay before display Er:29 (AFS E Alarm Elix product resistivity<SP 1 min UV HF) or (ELIX) Er:11 Er:12 Er:13 Er:14 Alarm Elix temp<0°C 10 sec Alarm Alarm Elix temp>50°C Elix res<0,2 MΩ.cm (@25°C Elix res>85 MΩ.cm (@25°C 10 sec 10 sec Alarm 10 sec Continued on next page SM_ESSENTIAL RANGE_TC&EC 91-172 EDI Resistivity Cell – EC006, Continued Test of the EDI Resistivity & thermistor Circuit Test of the EDI Resistivity & thermistor Cell The test must be carried out with the diagnostic box as indicated below : Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the RO/Feed conductivity circuit by using the corresponding rotary switch on the diagnostic box. • • Step 1 2 3 4 5 Action Visual inspection of the 2 resistivity electrodes. See Main Control PC Board EC002 to identify the EDI Resistivity cell pins on harness assembly J12 and perform a continuity test. Test the EDI Resistivity Cell by measuring the resistance between the 2 wires on the connector while disconnected from the Main board. The measured resistance should be very high. The measured resistance will climb during the time the Voltmeter is used. Perform test with a resistor in place of the thermistor (100kΩ = 25°C) Perform test with a resistor in place of the resistivity cell to test the cell wires and the signal acquisition by the Main board. Step Action 1 Power off the system. 2 Disconnect the conductivity cell connector. 3 Knowing that the EDI Resistivity cell constant is equal to 0.03 cm-1, a resistor of 450 KΩ is similar to a resistivity of 15 MΩ.cm. Continued on next page SM_ESSENTIAL RANGE_TC&EC 92-172 EDI Resistivity Cell – EC006, Continued Test of the EDI Resistivity & thermistor Cell, Continued Step 5 Step 4 5 SM_ESSENTIAL RANGE_TC&EC Action Action Connect the resistor between the electrodes Power on the system and verify the measured value with Millitrack. 93-172 Distribution Resistivity Cell – EC007 Description The Distribution Resistivity Cell is a sensor that measures the product water resistivity. The resistivity cell includes a thermistor in order to measure the temperature of the water. The temperature of the product is used to compensate the product resistivity. Location The Distribution Resistivity cell is used in AFS D/AFS HF/AFS-E HF AFS D Continued on next page SM_ESSENTIAL RANGE_TC&EC 94-172 Distribution Resistivity Cell – EC007, Continued Location AFS E HF Continued on next page SM_ESSENTIAL RANGE_TC&EC 95-172 Distribution Resistivity Cell – EC007, Continued Location AFS UV HF Catalog Number The Distribution (Product) Resistivity Cell catalog number is ZF3000518 for AFS-D (6/6mm). The Distribution (Product) Resistivity Cell catalog number is ZF3000519 for AFS UV HF/AFS-E UV HF (8/8mm). Continued on next page SM_ESSENTIAL RANGE_TC&EC 96-172 Distribution Resistivity Cell – EC007, Continued Structure The picture below shows the structure of the Elix Resistivity Cell: 4 1 2 5 3 6 Not used 1 Number Part Function 1 Resistivity cell wires 2 Thermistor wire 3 Water outlet (port size: 6 or 8 mm) Built-in Strainer Measures the resistivity of the water after the Elix module. Measures the temperature of the water after the Elix module. Water goes out 4 5 6 Not shown Water inlet (port size: 6 or 8 mm) O-Ring Check valve Protects both the cell and the flowmeter down stream. Strainer pore size = 190 µm Water goes in Catalog Number ZF3000518 or ZF3000519 Prevent leaks Prevent water from the tank to go in the Elix module Characteristics The table below describes the characteristics of the Distribution Resistivity Cell : Cell constant 0.03 cm-1 Thermistor 100 KΩ @25°C Port sizes 6/6 mm or 8/8 mm Continued on next page SM_ESSENTIAL RANGE_TC&EC 97-172 Distribution Resistivity Cell – EC007, Continued Software Interaction The table below describes software interactions of the cell : Description Distribution Resistivity Display The Distribution Resistivity is shown in the Millitrack Advanced parameters tab (Measures). Define the Distribution In the Millitrack Configuration tab, the setpoint Resistivity setpoint value of the Distribution Resistiv can be value; when the adjusted. Distribution Resistivity is above this value, the system displays a major alert. Diagnostic A problem with the Distribution Resistivity Cell can cause the following symptoms : Message Er:15 Er:16 Type What it means Alarm Distribution Resistivity<SP Distribution temp<0°C Distribution temp>50°C Distribution res<0,5 MΩ.cm@25°C Distribution res>85 MΩ.cm@25°C Alarm Alarm Er:17 Alarm Er:18 Alarm SM_ESSENTIAL RANGE_TC&EC Delay before display 40 sec 10 sec 10 sec 10 sec 10 sec 98-172 Distribution Resistivity Cell – EC007, Continued Test of the Distribution Resistivity circuit Test of the Distribution Resistivity & Thermistor Cell The test must be carried out with the diagnostic box as indicated below : Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the Distribution resistivity circuit by using the corresponding rotary switch on the diagnostic box. • • Step 1 2 3 4 Action Visual inspection of the 2 electrodes. See Main Control PC Board EC002 to identify the Distribution Resistivity cell pins on harness assembly J12 and perform a continuity test. Test the Distribution Resistivity Cell by measuring the resistance between the 2 wires on the connector while disconnected from the Main board. The measured resistance should be very high. The measured resistance will climb during the time the Voltmeter is used. Perform test with a resistor in place of the thermistor (100kΩ = 25°C) Continued on next page SM_ESSENTIAL RANGE_TC&EC 99-172 Distribution Resistivity Cell – EC007, Continued Test of the Distribution Resistivity & Thermistor Cell (continued) Step 5 Action Perform test with a resistor in place of the resistivity cell to test the cell wires and the signal acquisition by the Main board. Step Action 1 Power off the system. 2 Disconnect the Resistivity cell connector. 3 4 5 SM_ESSENTIAL RANGE_TC&EC Knowing that the Distribution Resistivity cell constant is equal to 0.03 cm-1, a resistor of 450 KΩ is similar to a resistivity of 15 MΩ.cm. Connect the resistor between the electrodes. Power on the system and verify the measured value with Millitrack. 100-172 RO Pressure Sensor – EC008 Description The Pressure Sensor measures the pressure in the system after the RO pump. The Pressure Sensor (PS) is pressurised by the RO Reject Water. The PS is an analogue circuit. The PS is fed with 5V and sends back a signal of ~0,3Volt for every Bar of water pressure. Location The RO Pressure sensor is located on the manifold as shown below : ELIX Continued on next page SM_ESSENTIAL RANGE_TC&EC 101-172 RO Pressure Sensor – EC008, Continued Location AFS E UV HF Catalog Number The RO Pressure Sensor catalog number is ZF3000520 : ACM_PRESSURE SENSOR 0-14 BARS Structure Continued on next page SM_ESSENTIAL RANGE_TC&EC 102-172 RO Pressure Sensor – EC008, Continued Characteristics The Pressure Sensor measurement range is 0-14 bars. Software Interaction The table below describes software interactions of the sensor : Diagnostic A faulty Pressure Sensor can cause a “Low Feed Pressure” alarm Description Display the RO Pressure Display The RO Pressure is shown in the Millitrack Advanced parameters tab (Measures). displayed by the icon The displayed pressure would be 0 Bar. Test of the RO Pressure circuit The test must be carried out with the diagnostic box as indicated below : • • Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the RO Pressure circuit by using the corresponding rotary switch on the diagnostic box. Continued on next page SM_ESSENTIAL RANGE_TC&EC 103-172 RO Pressure Sensor – EC008, Continued Test of the RO Pressure Sensor Step 1 2 3 4 5 6 7 8 Action Put the system in standby Visual inspection of the connection See Main Control PC Board EC002 to identify the RO Pressure sensor pins on harness assembly J12. Perform a continuity test (harness assembly) Remove and disconnect sensor Measure voltage (5V) on the connector to make sure the sensor is fed with 5V Reconnect the sensor Apply a small pressure directly on the membrane of the pressure sensor just by slightly pressing on the membrane with a flat tool. 9 Check the pressure measured in the dashboard or in the Advanced Parameters Measures tab 10 If no pressure variation is detected, replace the pressure sensor SM_ESSENTIAL RANGE_TC&EC 104-172 Distribution Pressure sensor – EC009 Description The pressure sensor measures the pressure in the distribution loop and allows to generate an Alarm stop (Er:31) if the pressure exceeds 6 bar. Location The distribution Pressure sensor is used in AFS-D/AFS-D UV/AFS HF/AFS-E HF AFSD UV Catalog Number The Distribution Pressure sensor catalog number is : ZF3000520 : ACM_PRESSURE SENSOR 0-14 BARS Continued on next page SM_ESSENTIAL RANGE_TC&EC 105-172 Distribution Pressure sensor – EC009, Continued Structure Characteristics The Pressure Sensor measurement range is 0-14 bars. Software Interaction Diagnostic Test of the Distribution Pressure circuit Description Distribution Pressure display Display The Distribution Pressure information is shown in the Millitrack Advanced parameters tab (Measures). A problem with the Distribution Pressure Sensor can cause the following symptoms : Message Type What it means Er:23 Alarm Stop Er:31 Alarm Stop Pressure sensor failure/sensor unplugged • Distribution loop overpressure • Recirculation loop malfunction (recirculation or relief valve) • Real overpressure Delay before display 1 sec 1 sec The test must be carried out with the diagnostic box as indicated below : • • Plug the diagnostic box on the system via the Main Board harness J12 as described in EC001. Test the Distribution Pressure sensor circuit by using the corresponding rotary switch on the diagnostic box. Continued on next page SM_ESSENTIAL RANGE_TC&EC 106-172 Distribution Pressure sensor – EC009, Continued Test of the Distribution Pressure Sensor Step 1 2 3 4 5 6 7 8 Action Put the system in standby Visual inspection of the connection See Main Control PC Board EC002 to identify the Distribution Pressure sensor pins on harness assembly J12. Perform a continuity test (harness assembly) Remove and disconnect sensor Measure voltage (5V) on the connector to make sure the sensor is fed with 5V Reconnect the sensor Apply a small pressure directly on the membrane of the pressure sensor just by slightly pressing on the membrane with a flat tool. 9 Check the pressure measured in the dashboard or in the Advanced Parameters Measures tab 10 If no pressure variation is detected, replace the pressure sensor SM_ESSENTIAL RANGE_TC&EC 107-172 RO Inlet Solenoid Valve – EC010 Description The RO Inlet Solenoid Valve: • controls feed water entering into the system, • is a normally closed 2-way solenoid valve (closed when not powered). Location Continued on next page SM_ESSENTIAL RANGE_TC&EC 108-172 RO Inlet Solenoid Valve – EC010, Continued Catalog Number SOLENOID VALVE MANIF. 2 WAY: PF05067 ACM_SOLENOID COIL (24 VDC) : ZF3000522 Spare Parts The picture below describes the structure of a solenoid valve: 4 2 1 2 1 2a 3 2b 4 2c 5 Number 1 2 2a 2b 2c 3 4 5 Part Valve o-ring Solenoid valve 2-way Body Spring Spring Electrical coil (24 VDC) Solenoid valve nut Torx 10 screw Spare Part FTPF05070 FTPF05067 ZF3000522 FTPF07915 Continued on next page SM_ESSENTIAL RANGE_TC&EC 109-172 RO Inlet Solenoid Valve – EC010, Continued Parker Coil Characteristics The picture shows the signal applied to the solenoid valve coils 24 Vdc By using a duty cycle (energy saving mode), the Average Operating Voltage is lowered to about 11 VDC. A voltmeter is not able to display such a signal. Continued on next page SM_ESSENTIAL RANGE_TC&EC 110-172 RO Inlet Solenoid Valve – EC010, Continued Electrical Connection Zoom in J2 Main Control PC Board J2 Software Interaction The RO Inlet Solenoid valve can be tested via the Millitrack dashboard. Continued on next page SM_ESSENTIAL RANGE_TC&EC 111-172 RO Inlet Solenoid Valve – EC010, Continued Diagnostic A problem with the RO Inlet Solenoid Valve can cause the following symptom: No Elix product flow, Low pressure alarm. The list above is not exhaustive. Check the Inlet valve circuit and the RO Inlet Valve The Inlet valve can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press Inlet valve When tested and electrically powered, the Inlet valve is highlighted in yellow in the dashboard as shown above. Continued on next page SM_ESSENTIAL RANGE_TC&EC 112-172 RO Inlet Solenoid Valve – EC010, Continued Test of the RO Inlet Solenoid Valve • Measure the impedance of the solenoid valve by measuring the resistance between the 2 wires. It should be around 35 Ω. • To test the RO Inlet Solenoid Valve, proceed as follows : Step 1 2 3 4 5 6 7 8 9 Action Power off the system. Open the left side cover. Inside the system: disconnect the tubing connected to the RO Inlet Solenoid Valve. Connect a tubing to the RO Inlet Solenoid Valve and put the other side of this tubing to a sink. Check that no water is pouring out of this tubing. Power on the system. Wait until the system is in flush mode. Check that water is coming out of this tubing. Analysis: If the test in steps 5 and 8 are… 10 Then the RO Inlet Solenoid Valve is … OK. OK. not OK. faulty. Re-connect the RO Inlet Solenoid Valve to the system and screw back the cover. SM_ESSENTIAL RANGE_TC&EC 113-172 RO Flush Solenoid Valve – EC011 Description The RO Flush Solenoid Valve is powered during RO flush. The RO Flush Solenoid Valve is a normally closed 2-way solenoid valve (closed when not powered). It is open during the Flush mode. Location Continued on next page SM_ESSENTIAL RANGE_TC&EC 114-172 RO Flush Solenoid Valve – EC011, Continued Catalog Number SOLENOID VALVE MANIF. 2 WAY: PF05067 ACM_SOLENOID COIL (24 VDC) : ZF3000522 Spare Parts The picture below describes the structure of a solenoid valve : 4 2 1 2 1 2a 3 2b 4 2c 5 Number 1 2 2a 2b 2c 3 4 5 Part Valve o-ring Solenoid valve 2-way Body Spring Spring Electrical coil (24 VDC) Solenoid valve nut Torx 10 screw Spare Part FTPF05070 FTPF05067 ZF3000522 FTPF07915 Continued on next page SM_ESSENTIAL RANGE_TC&EC 115-172 RO Flush Solenoid Valve – EC011, Continued Parker Coil Characteristics The picture shows the signal applied to the solenoid valve coils: 24 Vdc By using a duty cycle (energy saving mode), the Average Operating Voltage is lowered to about 11 VDC. A voltmeter is not able to display such a signal. Continued on next page SM_ESSENTIAL RANGE_TC&EC 116-172 RO Flush Solenoid Valve – EC011, Continued Electrical Connection Zoom in J3 Main Control PC Board J3 Continued on next page SM_ESSENTIAL RANGE_TC&EC 117-172 RO Flush Solenoid Valve – EC011, Continued Software Interaction The RO Flush Solenoid valve can be tested via the Millitrack dashboard. Diagnostic A faulty RO Flush Solenoid Valve can shorten the RO membrane lifetime. Check the Flush Valve circuit and the RO Flush Valve The Flush valve circuit can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press Flush valve When tested and electrically powered, the Flush valve is highlighted in yellow in the dashboard as shown above. Continued on next page SM_ESSENTIAL RANGE_TC&EC 118-172 RO Flush Solenoid Valve – EC011, Continued Test of the RO Flush Solenoid Valve • Measure the impedance of the solenoid valve by measuring the resistance between the 2 wires. It should be around 35 Ω. • To test the RO Flush Solenoid Valve, proceed as follows : Step 1 2 3 4 5 6 7 8 9 10 Action Power off the system. Open the left side cover. Inside the system: disconnect the tubing connected to the RO Flush Solenoid Valve. Connect a tubing to the RO Flush Solenoid Valve and put the other side of this tubing to a sink. Check that no water is pouring out of this tubing. Power on the system. Wait until the system is in flush mode. Check that water is coming out of this tubing. Analysis: If the test in steps 5 and 8 Then the RO Flush are… Solenoid Valve is … OK. OK. not OK. faulty. Re-connect the RO Flush Solenoid Valve to the system and screw back the cover. SM_ESSENTIAL RANGE_TC&EC 119-172 RO Permeate Divert Solenoid Valve – EC012 Description The RO Permeate 3-way Solenoid Valve controls water going from the RO permeate to: • the EDI module inlet during TANK FILLING mode (Normally open path), • the RO reject during RINSING mode (when solenoid valve powered). Location Continued on next page SM_ESSENTIAL RANGE_TC&EC 120-172 RO Permeate Divert Solenoid Valve – EC012, Continued Catalog Number SOLENOID VALVE MANIF. 3 WAYS: PF05068 ACM_SOLENOID COIL (24 VDC) : ZF3000522 Characteristics The picture shows the signal applied to the solenoid valve coils: 24 Vdc By using a duty cycle, the Average Operating Voltage is lowered to about 11 VDC. A voltmeter is not able to display such a signal. Continued on next page SM_ESSENTIAL RANGE_TC&EC 121-172 RO Permeate Divert Solenoid Valve – EC012, Continued Electrical Connection Zoom in J4 Main Control PC Board J4 Continued on next page SM_ESSENTIAL RANGE_TC&EC 122-172 RO Permeate Divert Solenoid Valve – EC012, Continued Software Interaction The RO Permeate Divert Solenoid valve can be tested via the Millitrack dashboard. Diagnostic A faulty RO Permeate Solenoid Valve can shorten the RO membrane(s) and/or the EDI module lifetime. Check the RO Permeate Solenoid Valve (Rinsing valve) circuit and the RO Permeate Valve The RO Permeate Valve circuit can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press Rinsing valve When tested and electrically powered, the RO Permeate Valve is highlighted in yellow in the dashboard as shown above. Continued on next page SM_ESSENTIAL RANGE_TC&EC 123-172 RO Permeate Divert Solenoid Valve – EC012, Continued Test of the RO Permeate Divert Solenoid Valve • Measure the impedance of the solenoid valve by measuring the resistance between the 2 wires. It should be around 35 Ω. • To test the RO Permeate Divert Solenoid Valve, proceed as follows: Step 1 2 3 4 5 6 7 8 9 Action Power off the system. Open the left side cover. Inside the system: disconnect the tubing connected to the RO Inlet Solenoid Valve. Connect a tubing to the RO Permeate Divert Solenoid Valve and put the other side of this tubing to a sink. Check that no water is pouring out of this tubing. Power on the system. Wait until the system is in flush mode. Check that water is coming out of this tubing. Analysis: If the test in steps 5 and 8 are… 10 11 Then the RO Permeate Divert Solenoid Valve is … OK. OK. not OK. faulty. Re-connect the RO Permeate Divert Solenoid Valve to the system and screw back the cover. Return to the Troubleshooting Card. SM_ESSENTIAL RANGE_TC&EC 124-172 Recirculation Solenoid Valve – EC013 Description The Recirculation 2-ways Solenoid Valve: • controls: - the flow rate in dispensing mode, the recirculation mode, • is a normally closed 2-ways solenoid valve (closed when not powered). Location The Recirculation Solenoid Valve is used on AFS-D & AFS- D UV. AFS-D UV Continued on next page SM_ESSENTIAL RANGE_TC&EC 125-172 Recirculation Solenoid Valve – EC013, Continued Catalog Number ACM_AFS D RECIRC EV COIL : ZF3000523 Structure The picture below describes the structure of a solenoid valve: 4 2 1 2 1 2a 3 2b 4 2c 5 Number 1 2 2a 2b 2c 3 4 5 Part Valve o-ring Solenoid valve 2-way Body Spring Spring Electrical coil (25 VDC) Solenoid valve nut Torx 10 screw Spare Part FTPF07918 FTPF09752 ZF3000523 FTPF07915 Characteristics The picture shows the signal applied to the solenoid valve coils: By using a duty cycle, the Average Operating Voltage is lowered to about 11 VDC. A voltmeter is not able to display such a signal. Continued on next page SM_ESSENTIAL RANGE_TC&EC 126-172 Recirculation Solenoid Valve – EC013, Continued Electrical Connection Zoom in J5 Main Control PC Board J5 Continued on next page SM_ESSENTIAL RANGE_TC&EC 127-172 Recirculation Solenoid Valve – EC013, Continued Software Interaction The Recirculation Solenoid valve can be tested via the Millitrack dashboard. Diagnostic A problem with the Recirculation Solenoid Valve can cause the following symptoms: • Low product flow, • Low Flow Dispensing not possible, • Bad water quality because recirculation is not possible. The list above is not exhaustive. Check the Recirculation Solenoid Valve circuit and the recirculation Valve The Recirculation Valve circuit can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press Recirculation Valve When tested and electrically powered, the Recirculation Valve is highlighted in yellow in the dashboard as shown above. Continued on next page SM_ESSENTIAL RANGE_TC&EC 128-172 Recirculation Solenoid Valve – EC013, Continued Test of the Recirculation Solenoid Valve • Measure the impedance of the solenoid valve by measuring the resistance between the 2 wires. It should be around 50 Ω. • To test the Recirculation Solenoid Valve, proceed as follows : Step 1 2 3 4 5 6 7 8 9 10 11 Action Power off the system. Open the left side cover. Inside the system: disconnect the tubing connected to the Rerciculation Solenoid Valve. Connect a tubing to the Recirculation Solenoid Valve and put the other side of this tubing to a sink. Check that no water is pouring out of this tubing. Power on the system. Wait until the system is in flush mode. Check that water is coming out of this tubing. Analysis: If the test in steps 5 and 8 Then the Rercirculation are… Solenoid Valve is … OK. OK. not OK. faulty. Re-connect the Rercirculation Solenoid Valve to the system and screw back the cover. Return to the Troubleshooting Card. SM_ESSENTIAL RANGE_TC&EC 129-172 254nm UV Lamp – EC014 Description ! Hazard Location The 254nm UV Lamp ensures bacteria destruction. The light from a UV Lamp is very damaging to eyes and skin. Do not turn on the Water System or attempt to turn on the UV Lamp in any way until the UV Lamp is inside its Stainless Steel Housing inside the Water System. The 254nm UV Lamp is used on AFS-D UV/AFS UV HF/AFSE UV HF/ELIX UV. AFSD UV Continued on next page SM_ESSENTIAL RANGE_TC&EC 130-172 254nm UV Lamp – EC014, Continued Location AFSE UV HF Catalog Number ACM_UV HOUSING ASSEMBLY ELIX : ZF3000525 ACM_ UV HOUSING ASSEMBLY AFS : ZF3000526 Continued on next page SM_ESSENTIAL RANGE_TC&EC 131-172 254nm UV Lamp – EC014, Continued Structure of the UV Lamp The picture below describes the structure of the UV lamp assembly : 1 2 3 The table below describes the function of each part : Part 1 2 3 Not shown Description Connection to the UV Ballast board 254 nm UV Lamp UV Lamp Housing 8 mm tubing connector Catalog Number FTPF08926 ZLXUVLP01 ZF3000525 (Elix)/ZF3000526 (AFS) Characteristics The characteristics of the UV lamp and housing are described below: • 10 Watt UV Lamp (AFS systems), • 6 Watt UV Lamp (Elix UV), • 8 mm Tefen Ports. Continued on next page SM_ESSENTIAL RANGE_TC&EC 132-172 254nm UV Lamp – EC014, Continued Electrical Connection UV BALLAST PCB Pay attention to the microswitches setting ! 10 W Configuration for: • AFS D • AFS UV HF • AFS E UV HF 6W Configuration for : • ELIX UV Continued on next page SM_ESSENTIAL RANGE_TC&EC 133-172 254nm UV Lamp – EC014, Continued Electrical Connection UV Ballast Board Main Control PCB J7 Software Interaction The status and current level of the UV Ballast is given in the Advanced Parameters tab (Hardware status). Continued on next page SM_ESSENTIAL RANGE_TC&EC 134-172 254nm UV Lamp – EC014, Continued Diagnostic A problem with the UV Lamp can cause the alert message: Error code 25 (Check UV Lamp) It means that the intensity coming from the UV Ballast is below 147 mA. Other issues can also appear. Test of Main board, UV Ballast board and 254 nm UV Lamp The 254nm UV lamp circuit and the UV Lamp can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press 254nm UV lamp When tested and electrically powered, the 254nm UV Lamp is highlighted in the dashboard as shown above. The 254nm UV Lamp circuit can be tested by plugging the 10W resistor on the Main Board (J7) (See EC001 Tools needed to perform system test) Continued on next page SM_ESSENTIAL RANGE_TC&EC 135-172 254nm UV Lamp – EC014, Continued To test the 254 nm UV Lamp, proceed as follows: Test of Main board, UV Step Action Ballast board 0 Make sure that the UV 254 nm Lamp is not deactivated in the and 254 nm UV CONFIG Software. Lamp 1 Power off the system. (continued) 2 Replace the UV Lamp by a bulb with 4 pins as the one shown below: 3 4 5 6 Note: Use a ~18-25 Watts lamp with the same connector as the 254 nm UV Lamp. Power on the system. Start a recirculation. Wait 10 seconds. Result: If the bulb... turns on. does not turn on. SM_ESSENTIAL RANGE_TC&EC Then ... The 254 nm UV Lamp is faulty. UV Ballast board is faulty. 136-172 254nm UV Lamp – EC014, Continued Correct Connection When installing a UV Lamp in a water system, make sure that the cable is connected as shown below. This is recommended to avoid potential lamp ignition problems. The black and white cable wires must be connected on the same side as the white wires inside the lamp. SM_ESSENTIAL RANGE_TC&EC 137-172 Distribution Pump High Flowrate – EC015 Description The Distribution Pump allows dispensing and recirculation. Location The distribution Pump High Flowrate is fitted in the underbox on AFS (UV) HF & AFS-E (UV) HF systems. The location of the Distribution Pump is shown below: AFS (UV) HF/AFS-E (UV) HF Continued on next page SM_ESSENTIAL RANGE_TC&EC 138-172 Distribution Pump High Flowrate – EC015, Continued Catalog Number ACM_DIST_DIAPHRAGM PUMP (AQU8872) : ZF3000521 Structure The picture below describes the structure of the Distribution Pump: 1 4 2 3 The table below describes the function of each part : Part 1 2 3 4 Characteristics Description Elbow fitting 8M-8T Distribution pump Ferrite Supply wires Catalog Number FTPF05143 ZF3000521 The table below shows the characteristics of the Distribution Pump: Voltage (max) Type of connector Flow Rate ~24 VDC Molex 2 pins 2.2 lpm Continued on next page SM_ESSENTIAL RANGE_TC&EC 139-172 Distribution Pump High Flowrate – EC015, Continued Electrical Connection Connect the Distribution pump to the J8 connector as shown below : J8 Software Interaction The status and current level of the Distribution Pump are given in the Advanced Parameters tab (Hardware status). Continued on next page SM_ESSENTIAL RANGE_TC&EC 140-172 Distribution Pump High Flowrate – EC015, Continued Diagnostic A problem with the Distribution Pump can cause the following symptoms: • No product flow, • Low product flow. The list above is not exhaustive. Test of the J8 The J8 circuit and the distribution pump can be electrically tested by Circuit and the means of the Millitrack dashboard. distribution Follow the procedure below described : pump : The system must be in Standby mode to enable components tests. • Press Enable tests • Press Distribution pump When tested and electrically powered, the voltage and current to the pump are displayed. Continued on next page SM_ESSENTIAL RANGE_TC&EC 141-172 Distribution Pump High Flowrate – EC015, Continued Test of the Distribution Pump To test the Distribution Pump, proceed as follows : Check the voltage and current that are displayed in the Millitrack dashboard: • If 0 A is displayed in Distribution operation, the Distribution pump could be defective. Connect the UV Resistor (50Ω) used to simulate the UV Lamp in place of the distribution Pump on the Main board J8 connector. If a current value is displayed, it confirms that the Distribution pump could be defective and needs to be replaced. • If a current value is displayed but the voltage remains at 0V in Distribution operation, the Main board is probably defective and needs to be replaced. Unplug the Distribution Pump and check the presence of the 24V on the Main board J8 connector during distribution mode. SM_ESSENTIAL RANGE_TC&EC 142-172 Distribution Pump Low Flowrate – EC016 Description The low flowrate Distribution Pump and vacuum Pump allows dispensing and recirculation. Location The distribution pump low flowrate is fitted on AFS-D & AFS-D UV. The location of the Distribution Pump is shown below : AFSD/AFSD UV Continued on next page SM_ESSENTIAL RANGE_TC&EC 143-172 Distribution Pump Low Flowrate – EC016, Continued Catalog Number WPM DIST/VAC PUMP 24 VDC (PUMP BRUSHLESS) : ZF3000272 Structure The picture below describes the structure of the Distribution Pump: Distribution and vacuum pump/Dual pump head OUT IN IN OUT Characteristics The table below shows the characteristics of the Distribution Pump: Voltage (max) Type of connector Flow Rate Electrical Connection ~24 VDC Molex 2 pins 0.6 lpm Connect the Distribution pump to the J8 connector as shown below : J8 Continued on next page SM_ESSENTIAL RANGE_TC&EC 144-172 Distribution Pump Low Flowrate – EC016, Continued Software Interaction The status and current level of the Distribution Pump are given in the Advanced Parameters tab (Hardware status). Diagnostic A problem with the Distribution Pump can cause the following symptoms: • No product flow, • Low product flow. The list above is not exhaustive. Test of the J8 Circuit The J8 circuit can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press Distribution pump When tested and electrically powered, the voltage and current to the pump are displayed. Continued on next page SM_ESSENTIAL RANGE_TC&EC 145-172 Distribution Pump Low Flowrate – EC016, Continued Test of the Distribution Pump To test the Distribution Pump, proceed as follows : Check the voltage and current that are displayed in the Millitrack dashboard. • • If 0 A is displayed in Distribution operation, the Distribution pump could be defective. Connect the 10W Resistor used to simulate the UV Lamp in place of the distribution Pump on the Main board J8 connector. If a current value is displayed, it confirms that the Distribution pump is defective and needs to be replaced. If a current value is displayed but the voltage remains at 0V in Distribution operation, the Main board is probably defective and needs to be replaced. Unplug the Distribution Pump and check the presence of the 24V on the Main board J8 connector during distribution operation. SM_ESSENTIAL RANGE_TC&EC 146-172 RO Pump – EC017 Description The RO Pump allows: • The RO membrane to be fed with the correct pressure (depending on feed water temperature), • The Elix part of the system to run, The RO Pump is supplied by the Main PCB with a voltage in a range [+12V, + 40V (1.3A)] Location The RO pump is fitted on RIOS/ELIX & AFS systems. RIOS Continued on next page SM_ESSENTIAL RANGE_TC&EC 147-172 RO Pump – EC017, Continued Location, continued Catalog Number AFSD UV ACM_DIAPHRAGM PUMP AQU8871 : ZF3000528 Continued on next page SM_ESSENTIAL RANGE_TC&EC 148-172 RO Pump – EC017, Continued Structure The picture below describes the structure of the RO Pump: 1 2 3 4 Part 1 2 3 4 Description Elbow fitting 8M-8T RO pump Ferrite Supply wires Catalog Number FTPF05143 ZF3000528 Characteristics The table below shows the characteristics of the RO Pump: Voltage (max) Type of connector ~40 VDC Molex 3 pins Continued on next page SM_ESSENTIAL RANGE_TC&EC 149-172 RO Pump – EC017, Continued Electrical Connection Connect the RO pump to the J9 3 pins Molex connector : J9 Software Interaction The status, the current level and the consign of the RO Pump are given in the Millitrack Advanced Parameters tab (Hardware status). Diagnostic A problem with the RO Pump can cause the following symptoms: • Low feed pressure alarm, • Low Elix product flow : the tank is not filled. The list above is not exhaustive. Continued on next page SM_ESSENTIAL RANGE_TC&EC 150-172 RO Pump – EC017, Continued Test of the J9 Circuit To test the J9 Circuit proceed as follows: Step 1 2 3 4 Action Power off the water system. Remove the Progard Pack and place plugs instead. Disconnect the Distribution Pump from the J9 connector of the Main Control PCB. Place a voltmeter with parrot clip between pin 1 and 3 as shown below: V 5 6 Power on the system. Go in Millitrack and put the RO Pump speed in manual mode. 7 8 Modify the RO Pump speed and check the voltage. Analysis: If ... • Voltage variation between 0V and 40V with the RO Pump Speed Adjustment • No voltage variation between 0V and 40V with the RO Pump Speed Adjustment Then the J9 Circuit is... OK faulty Continued on next page SM_ESSENTIAL RANGE_TC&EC 151-172 RO Pump – EC017, Continued Test of the RO Pump To test the RO Pump, proceed as follows : Check the voltage and current that are displayed in the Millitrack dashboard when the RO Pump is powered (Rinsing, Filling tank). • If 0 A is displayed, the RO Pump could be defective • If 0V is displayed, the electronic circuit could be defective. Perform circuit test as described in the previous page. Try to force the motor speed by selecting the Pump speed manual activation and check current and voltage variation according to the Pump speed adjustment. SM_ESSENTIAL RANGE_TC&EC 152-172 MAIN DISPLAY – EC018 Description The Main Display consists of the following items: LCD, Backlight. MMI (Man-Machine Interface) Display Location The Main Display is located as shown below : Catalog Number ACM LCD DISPLAY : ZF3000509 SM_ESSENTIAL RANGE_TC&EC 153-172 MAIN DISPLAY – EC018, Continued Characteristics The Main Display is a LCD with a LED type backlight. Electrical Connection The LCD is connected to the “LCD” connector of the Main Control PCB has shown in blue below. J13 Software Interaction LCD ECO MODE When this function is activated, the ECO mode is used to save energy when the display is not used. It shuts down the display backlight after a 15 min period of non-use. The ECO mode is disabled by pressing a keypad button or when a warning message is displayed by the system (error code or warning symbol). Continued on next page SM_ESSENTIAL RANGE_TC&EC 154-172 MAIN DISPLAY – EC018, Continued Diagnostic A problem with the Main Display cell can cause the following symptoms: •No Display, •Main Display is dark because there is no backlight. The list above is not exhaustive. SM_ESSENTIAL RANGE_TC&EC 155-172 RFID Reader – EC019 Description The RFID Reader allows to read the pack tags and to detect the presence of pack. Location Catalog Number ACM_TAG READER PCB : ZF3000508 Continued on next page SM_ESSENTIAL RANGE_TC&EC 156-172 RFID Reader – EC019, Continued Electrical Connection The tag reader PCBs are connected to the Main PCB through connectors J16 and J17. J16 : RFID Pretreatment J17 : RFID Polisher Software Interaction The Pretreatment and Polisher Reader versions are shown in the Millitrack system software information : Diagnostic If something is wrong with the tag reader PCBs, the MMI icons indicating Progard/Polisher pack is out or incorrect will be displayed. This could mean that the Pack tag has not been detected. Continued on next page SM_ESSENTIAL RANGE_TC&EC 157-172 RFID Reader – EC019, Continued Test of the RFID Reader The pretreatment or Polisher Reader Version should be different from 0. If the Pretreatment or Polisher Reader version is 0, then the tag reader is defective. Test of the RFID Reader with the tag diagnostic tool The tag diagnostic tool (Catalog Number : ZF3000461) also used for troubleshooting on the Integral systems allows to verify whether the tag reader is defective. SM_ESSENTIAL RANGE_TC&EC 158-172 EDI Module – EC020 Description Location Catalog Number The EDI module removes ions by electrical deionization. The EDI Module is supplied by the EDI power supply board which controls the intensity regulation in a voltage range of 0-140 V. The EDI Module is used in Elix/Elix UV/AFS-E HF. ZLXE003WW-ELIX ESSENTIAL 3 ZLXEV03WW-ELIX ESSENTIAL 3 UV ZLXE005WW-ELIX ESSENTIAL 5 ZLXEV05WW-ELIX ESSENTIAL 5 UV ZLXE010WW-ELIX ESSENTIAL 10 ZLXEV10WW-ELIX ESSENTIAL 10 UV ZLXE015WW-ELIX ESSENTIAL 15 ZLXEV15WW-ELIX ESSENTIAL 15 UV ZAFS10EWW-AFS-10E ZAFS15EWW-AFS-15E Continued on next page SM_ESSENTIAL RANGE_TC&EC 159-172 EDI Module – EC020, Continued Structure The picture below describes the structure of the EDI module: 1 Waste Inlet 3 2 4 Product Inlet 1 1 The table below describes the catalog number of each part: Part 1 2 3 4 - Description Elbow fitting 6M-6T EDI module Supply wires Ferrite 6 mm tubing and Y connector Catalog Number ZLXE003WW-ELIX ESSENTIAL 3 ZLXEV03WW-ELIX ESSENTIAL 3 UV ZLXE005WW-ELIX ESSENTIAL 5 ZLXEV05WW-ELIX ESSENTIAL 5 UV ZLXE010WW-ELIX ESSENTIAL 10 ZLXEV10WW-ELIX ESSENTIAL 10 UV ZLXE015WW-ELIX ESSENTIAL 15 ZLXEV15WW-ELIX ESSENTIAL 15 UV ZAFS10EWW -AFS-10E ZAFS15EWW -AFS-15E Continued on next page SM_ESSENTIAL RANGE_TC&EC 160-172 EDI Module – EC020, Continued Characteristics The tables below show the characteristics of the EDI module : Characteristic Voltage max Intensity range Type of connector Value 140 VDC 70 – 115 mA Molex 2 pins The EDI Board controls the EDI Modules with 3 different intensities as shown below : Flow Rate [L/h] Iout(mA) Elix Elix UV AFS E UV HF Electrical Connection 3 70 x x 5 70 x x 10 100 x x x 15 115 x x x Connect the black and red wires from the EDI Module to the EDI Board. J6 Pay attention to the jumper position! EDI Module SM_ESSENTIAL RANGE_TC&EC 161-172 EDI Module – EC020, Continued Electrical Connection (continued) ZLXE003WW-ELIX ESSENTIAL 3 ZLXEV03WW-ELIX ESSENTIAL 3 UV ZLXE005WW-ELIX ESSENTIAL 5 ZLXEV05WW-ELIX ESSENTIAL 5 UV Jumper position 2-3 on left (70 mA) ZLXE010WW ELIX ESSENTIAL 10 ZLXEV10WW ELIX ESSENTIAL 10 UV Jumper position 2-2 (100 mA) ZAFS10EWW - AFS-10E ZLXE015WW ELIX ESSENTIAL 15 ZLXEV15WW ELIX ESSENTIAL 15 UV ZAFS15EWW - AFS-15E Jumper position 3-3 (115 mA) Software Interaction The EDI Power Supply Current is displayed in the Millitrack (Hardware status). Diagnostic A problem with the EDI module can cause the following symptoms: • • message, System reboot, or (ELIX PRODUCT R < SP) Alarm The list above is not exhaustive. Continued on next page SM_ESSENTIAL RANGE_TC&EC 162-172 EDI Module – EC020, Continued Test of the Circuit To test the circuit, proceed as follows : Step 1 2 3 4 Action Power off the system. Remove the Packs. Put plastic caps in the Pack adapter to avoid water spillage. Disconnect the EDI module from the connector of the EDI Power Supply PC Board. Place the 150 Ω EDI load (see EC001 list of tools needed for system test) on the two pins of connector as shown below: R The resistor load will become very hot! Avoid touching the resistor! ! 5 6 7 Power on the system. When the system goes in Tank Filling mode, check the EDI current in the Millitrack dashboard or in Hardware status. Analysis (depending on the EDI PS jumper position) : If voltage is… Then… U ≃ 10,5 VDC I (K1 value) 70 mA ≃ 15 VDC 100 mA ≃ 17,25 VDC 115 mA > 140 VDC or ≃ 0 VDC EDI Power Supply PCB is OK. Replace EDI Power Supply PCB. Continued on next page SM_ESSENTIAL RANGE_TC&EC 163-172 EDI Module – EC020, Continued Test of the circuit The J6 circuit can be electrically tested by means of the Millitrack dashboard. Follow the procedure below described : : The system must be in Standby mode to enable components tests. • Press Enable tests • Press EDI module When tested and electrically powered, the current to the EDI is displayed. Test of the EDI Module To test the resistance of the EDI module, proceed as follows : Step 2 3 5 6 Action Power off the system. Remove the Packs and put plastic caps in the Pack adapter to avoid water spillage. Disconnect the EDI module from the connector of the EDI Power Supply PC Board. Measure the impedance of the EDI module by placing an ohmmeter between the two pins in the Molex connector as shown below: Ω 7 Note: the EDI module must be full of water. Analysis: If resistance is < 1 kΩ, then repair contact of the cables or replace the EDI module. SM_ESSENTIAL RANGE_TC&EC 164-172 Harness – EC021 Description The harness cable that is connected to J12 on the Main PCB allows to connect all the sensors to the Main board. J20 : harness Catalog Numbers Catalog Number FTPF11996 FTPF11997 FTPF11998 FTPF11999 J12 on the Main PCB Description ACM RIOS HARNESS ASSY ACM ELIX HARNESS ASSY ACM AFS-D-HF HARNESS ASSY ACM AFS-E-UV-HF HARNESS ASSY Continued on next page SM_ESSENTIAL RANGE_TC&EC 165-172 Harness – EC021, Continued Electrical connection The system type harness contains all cables and wires used between the main control board and the different components (conductivity/resistivity, temperature and pressure). On Main PCB side, a single connector J12 regroups all wires from the different system sensors. 4 harness references have been developed. The following figure gives the Catalog Numbers and also the naming convention of the Essential harness : FTPF11999 HARNESS AFS-E-HF FTPF11998 HARNESS AFS-D / AFS-HF FTPF11997 HARNESS ELIX FTPF11996 HARNESS RIOS Continued on next page SM_ESSENTIAL RANGE_TC&EC 166-172 Harness – EC021, Continued Harness connection description The following figure shows the most complete ACME harness : Connectors: Stocko Socket Molex Female crimp terminal Molex Male crimp terminal Tyco Female connector Molex 20 pins female connector Continued on next page SM_ESSENTIAL RANGE_TC&EC 167-172 Harness – EC021, Continued J20 connector pin out The following figure shows the J20 Main connector pinout : SM_ESSENTIAL RANGE_TC&EC 168-172 External Power Supply – EC022 Description The Essential range water systems are powered with an external power supply. The power supply allows to convert 100 VAC-230 VAC (50-60 Hz) into 24 VDC. Location The power supply for Essential range systems is located outside of the systems and allows to connect the system to the main. Catalog Number Power Supplies References Item Number Description RIOS 150 W FTPF11986 X 200 W FTPF11985 ACM EXT POWER SUPPLY MED ACM EXT POWER SUPPLY HIGH AFS D/AFS D UV X AFS UV HF AFS E UV HF X X The power supplies used according to the system types are the following : Power Supply150W Power Supply 200W IMPORTANT Elix/ Elix UV X RiOs Elix Elix UV x x AFS D AFS D UV x AFS UV HF AFS E UV HF x x The power supplies 150W are resettable but the power supplies 200W are not and can be security latched. For recovery (if security latch), the 200 W power supply needs to be unpowered during a few minutes. SM_ESSENTIAL RANGE_TC&EC 169-172 External Relay – EC023 Description An external relay can be plugged onto the RIOS and ELIX ESSENTIAL systems. The relay allows to monitor and report the alarms. Location The relay is located on top and at the back of the system as shown below. Catalog Number KITRELESS Electrical connection • The relay is connected to the Main Board J8 connector (2 pins with Molex) via the Main PCB to relay cable Main board PCB to relay cable Continued on next page SM_ESSENTIAL RANGE_TC&EC 170-172 External Relay – EC023, Continued Electrical connection, (continued) The relay is connected to the accessory connector at the back of the system via the relay to accessory connector cable shown below. Relay to accessory connector cable Software interaction To activate the relay function, follow the below procedure : Step 1 Action In the “CONFIGURATION” tab, select “Relay” in the options activation and “ASM UV or Relay Activation”. Continued on next page SM_ESSENTIAL RANGE_TC&EC 171-172 External Relay – EC023, Continued Software interaction (continued) Step 2 Action Scroll down this tab and search the “Relay Activation Settings” line, clic on the “SET” button. 3 In the External Relay Configuration, select the alarms/alerts to be monitored and save the configuration. SM_ESSENTIAL RANGE_TC&EC 172-172
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )