QUALITY ASSURANCE AND QUALITY CONTROL OF A SOIL TESTING LABORATORY Yuan-Wu Chen Fourth Division, Environmental Analysis Laboratory (EAL), Environmental Protection Administration (EPA), Zhongli District, Taoyuan City, Taiwan E-mail: ywchen@epa.gov.tw ABSTRACT Aiming for better environmental quality, Taiwan EPA launched in August 1987 and set a milestone in Taiwan’s environmental protection efforts. The Environmental Analysis Laboratory (EAL) established on January 10, 1990 to provide policymakers environmental data required. To enhance the credibility in quality assurance in environmental analysis, EAL participates actively in international laboratory accreditation. It received first laboratory accreditation from NATA (National Association of Testing Authorities, Australia) on January 31, 1995. The grand opening of the National Environment Analysis Building in November 1998 was a noteworthy milestone that signaled a new era of environmental testing and research. "Soil and Groundwater Pollution Remediation Act" was officially announced by the President on Feb 2, 2000. Soil and Groundwater Pollution Remediation Fund Management Board established in November 2001 based on the Act for management of the fund and conducting works for the promotion of soil and groundwater pollution remediation and prevention. Through several revision corrections, the current version of the Act contains 8 chapters and 57 articles. EPA of Taiwan government also promulgated relevant soil pollution control standards in 2001 in accordance with Article 6 of this Act. And according to Article 10 of this Act, the methods and quality control guidelines for soil contaminants analysis shall be determined by the EAL, and the EAL will also be in charge of granting accreditation to environmental analysis laboratories. Based on these statutory responsibilities, the EAL has published 91 soil standard testing methods, 8 relevant quality assurance guidelines and with ISO /IEC 17025 granting accreditation to 48 soil environmental testing laboratories. Laboratory quality manual of EAL also strictly followed this standard method, QA/QC guidelines and ISO /IEC 17025. We have made a briefing on some essential QA/QC requirements relevant to soil analysis. We wish to share and exchange our experiences with all experts in agricultural soil analysis laboratories participating in this seminar. Keywords: QA/QC, soil testing laboratory INTRODUCTION Aiming at better environmental quality, Taiwan EPA was launched in August 1987 and set a milestone in Taiwan’s environmental protection efforts. Environmental analysis is by all means the cornerstone of environmental policymaking. The formulation and execution of policies, regulations, and standards will rely on the basis of accurate data. Effective environmental protection policies cannot be formulated without precise and reliable environmental analysis results. High quality environmental analyses are critical for all aspects of environmental protection work. The Environmental Analysis Laboratory (EAL) was established on January 10, 1990 to provide policymakers with the environmental data required. EAL is divided into five divisions that cover tasks including (1) analysis planning and administration, (2) air and physical analysis, (3) inorganic analysis, (4) organic and ultra-trace analysis, (5) environmental biological analysis. Under the outstanding leadership of those pioneer directors of the institute, all staff members are dedicated to the improvement of environmental analysis technology, development of standard methods, upgrading of the quality of environmental analysis data. It will provide environmental laboratories from various sectors with technical guidance. Meanwhile, it is also a national level environmental analysis lab that is responsible for setting up environmental analysis standards. To enhance the credibility in quality assurance in environmental analysis, EAL participates actively in international laboratory accreditation. It received its first laboratory accreditation from NATA (National Association of Testing Authorities, Australia) on January 31, 1995. The grand opening of the National Environment Analysis Building in November 1998 was a noteworthy milestone that signaled a new era of environmental testing and research. EAL is engaged in enhancing its intimate cooperation with academic research institutes for improving efficiency and techniques. Enhancing our capacities in environmental analysis have been and will be always one of EAL’s major thrusts. Taiwan has encountered continual issues of soil and groundwater pollution since 1981 due to illegal disposal of industrial wastewater, wastes and emissions of air pollution. Several well-known cases such as those of Taiwan Petrochemical Development Corporation (TPDC), Coin Chemical Industrial Co., Ltd., Jili Chemical Corp., RCA, and numerous illegal waste disposal sites throughout the island have become the headlines in the past years. To address the issue of soil and groundwater pollution, the Environmental Protection Administration (EPA) embarked on establishing relevant regulations. After years of study and debate, "Soil and Groundwater Pollution Remediation Act" was officially announced by the President on Feb 2, 2000. Soil and Groundwater Pollution Remediation Fund Management Board was established in November 2001 based on the Act for management of the fund and conducting works for promotion of soil and groundwater pollution remediation and prevention. Through several revisions, the current version of the Act contains 8 chapters and 57 articles. EPA of Taiwan government also promulgated relevant soil pollution control standards in 2001 in accordance with Article 6 of this Act as shown in Table 1. Based on Article 10 of this Act, all methods and quality control guidelines for soil, sediment and groundwater contaminants analysis shall be established by the administration department in the central government, which implies the current status and responsibility of EAL. Also it is required that in the proceeding of all kinds of investigation or remediation works on bottom sediment, and groundwater pollution, all kinds of the analysis of soil and groundwater pollution are necessary to be provided or submitted and have to be contracted by labs approved by authorized department of the central government except when approved by the central competent authority. EAL is the department that is in charge of issuing permission licenses of contracted labs. Currently, EAL is responsible for setting up relevant soil standard testing methods and relevant quality assurance guideline. Also it will implement accreditation to all environmental testing laboratories following the international guideline ISO /IEC 17025 based on the statutory authorization of relevant acts. Table 1. Soil pollution control standards in Taiwan Category Target pollutant Regulated criteria arsenic(As) 60 mg/kg cadmium(Cd) chromium(Cr) copper(Cu) Heavy metal mercury(Hg) nickel(Ni) 400 mg/kg 200 mg/kg(Edible crop farmland) 20 mg/kg 5 mg/kg(Edible crop farmland) 200 mg/kg Benzene 2000 mg/kg 500 mg/kg(Edible crop farmland) 2000 mg/kg 600 mg/kg(Edible crop farmland) 5 mg/kg Carbon tetrachloride 5 mg/kg lead(Pb) zinc(Zn) Organic compound 20 mg/kg 5 mg/kg(Edible crop farmland) 250 mg/kg Category Target pollutant Regulated criteria Chloroform 1,2-Dichloroethane 100 mg/kg cis-1,2-Dichloroethylene trans-1,2-Dichloroethylene 50 mg/kg 1,2-Dichloropropane 1,2-Dichlorobenzene 0.5 mg/kg 100 mg/kg 1,3-Dichlorobenzene 3,3'-Dichlorobenzidine 100 mg/kg 2 mg/kg Ethylbenzene 250 mg/kg 500 mg/kg Hexachlorobenzene Pentachlorophenol Tetrachloroethylene Toluene Total petroleum hydrocarbons(TPH) Trichloroethylene 2,4,5-Trichlorophenol Other organic compound 200 mg/kg 10 mg/kg 500 mg/kg 1000 mg/kg 60 mg/kg 350 mg/kg 2,4,6-Trichlorophenol Vinyl chloride 40 mg/kg 10 mg/kg Xylenes 500 mg/kg 0.04 mg/kg Aldrin Chlordane Pesticide 8 mg/kg 7 mg/kg 0.5 mg/kg 3 mg/kg 4,4'-Dichlorodiphenyl-triichloroethane(DDT) Dieldrin 0.04 mg/kg Endrin Heptachlor 20 mg/kg 0.2 mg/kg Toxaphene Endosulfan 0.6 mg/kg 60 mg/kg Dioxins(PCDD/Fs) 1000 ng-TEQ/kg Polychlorinated biphenyls(PCBs) 0.09 mg/kg SOIL STANDARD METHODS OF TAIWAN EPA Based on the nature of the pollutants, standard methods of Taiwan EPA are classified into 10 categories as follows: noise testing, analysis of air, water, drinking water, soil, sediment, wastes, environmental agents, toxic chemicals, and biological testing. EAL has established 604 standard environmental methods up to now. Ninety-one (91) of the standard methods are relevant to soil sample analysis as shown in Table 2. EAL, laboratories of local environmental bureau and all accredited laboratories must follow these standard methods so as to ensure a consistent data quality. Table 2. Standard Methods of Taiwan EPA category Numbers of Standard Method Air Analysis Methods: Wastes Analysis Methods: Soil Analysis Methods: Wastes/Soil Common Methods: Drinking Water Treatment Agent Methods: Water Quality Analysis Methods: Toxic Chemicals Analysis Methods: Environmental Agents Analysis Methods: Environmental Bioanalytical Methods: 153 66 25 66 25 167 35 14 53 Total Numbers of Standard Method 604 Soil control pollutant corresponding standard method Among the 91 soil relevant methods established by the EAL, those which are related to the regulated target pollutants have been used most frequently. One of the major causes is that both EPA and local government pay more attention on the environmental survey, monitoring and inspection on these pollutants. The liable pollution producer of those controlled sites also needs to contract the authorized environmental analysis laboratory to analyze these targeted pollutants. Therefore, standard methods relevant to regulated target pollutants usually have good economic incentives in Taiwan. These methods are listed including their method numbers as shown in Table 3. Table 3. Soil target pollutant corresponding standard method Category Target pollutant Standard method Pb、Cd、Cu、Cr、Zn、Ni Aqua regia digestion method (NIEA S321.64B) Microwave assisted aqua regia digestion method (NIEA S301.60B)/ Flame atomic absorption spectrophotometry method (NIEA M111.01C) Inductively coupled plasma-atomic emission spectrometry method (NIEA M104.02C) Inductively coupled plasma - mass spectrometry method (NIEA M105.01B) As hydrogen arsenite atomic absorption spectrometric method (NIEA S310.64B) Heavy metal Hg Mercury by cold vapor atomic absorption spectrophotometry method(NIEA M317.03B) Mercury by thermal decomposition, amalgamation,and atomic absorption Category Target pollutant Standard method spectrophotometry Method (NIEA M318.01B) Volatile organic compounds (VOCs) Semi-volatile organic compounds (SVOCs) Benzene、Carbon tetrachloride、 Chloroform、1,2-Dichloroethane、 cis-1,2-Dichloroethylene、 trans-1,2-Dichloroethylene、 1,2-Dichloropropane、 1,2-Dichlorobenzene、 1,3-Dichlorobenzene、 Ethylbenzene、Toluene、 Tetrachloroethylene、 Trichloroethylene、Vinyl chloride、 Xylenes Closed-system purge-and-trap and extraction for volatile organics method (NIEA M155.01C) Volatile organic compounds using equilibrium headspace analysis method (NIEA M157.00C)/ Volatile organic compounds by gas chromatography/mass spectrometry (GC/MS) method (NIEA M711.04C) 3,3'-Dichlorobenzidine、 Hexachlorobenzene、 Pentachlorophenol、 2,4,5-Trichlorophenol、 2,4,6-Trichlorophenol Ultrasonic extraction method (NIEA M167.01C) Soxhlet extraction method (NIEA M165.01C) / Semivolatile organic compounds by gas chromatography/mass spectrometry (GC/MS): Method (NIEA M731.02C) C6-C9 Closed-system purge-and-trap and extraction for volatile organics method (NIEA M155.01C) Volatile organic compounds using equilibrium headspace analysis method (NIEA M157.00C) / Total petroleum hydrocarbons by gas chromatography/flame ionization detector (GC/FID)method (NIEA S703.62B) C10-C40 Ultrasonic extraction method (NIEA M167.01C) Soxhlet extraction method (NIEA M165.01C)/ chromatography/flame ionization detector (GC/FID)method (NIEA S703.62B) Total petroleum hydrocarbons (TPH) Organochlorine pesticides (OCPs) Aldrin、Chlordane、 4,4'-Dichlorodiphenyl-triichloroeth ane(DDT)、Dieldrin、Endrin、 Heptachlor、Toxaphene、 Endosulfan other organic compound Dioxins(PCDD/Fs) Ultrasonic extraction method (NIEA M167.01C) Soxhlet extraction method (NIEA M165.01C) Silica gel cleanup method (NIEA M183.01C)/ Organochlorine pesticides by gas chromatography method (NIEA M618.05C) Polychlorinated dibenzo-p-dioxins and dibenzofurans by high-resolution gas chromatography/high-resolution mass spectrometry (HRGC/HRMS) method (NIEA M801.13B) Category Target pollutant Standard method Polychlorinated dibenzo-p-dioxins and dibenzofurans by gas chromatography/tandam mass spectrometry (GC-MS/MS) method (NIEA M805.00B) other organic compound Polychlorinated biphenyls(PCBs) Ultrasonic extraction method (NIEA M167.01C) Soxhlet extraction. method (NIEA M165.01C) Silica gel cleanup method (NIEA M183.01C)/ Polychlorinated biphenyls (PCBs) by gas chromatography method (NIEA M619.04C) ACCREDITATION PROCESS FOR THE ENVIRONMENTAL ANALYSIS LABORATORIES EAL has been engaged in the accreditation of environmental analysis laboratories since 1987. With a view to increasing the transparency of the accreditation procedure, EAL issued the "Guideline for the Authorization and Management of Environmental Analysis Laboratories" in 1990. It has been revised and amended several times since then. On November 19, 1997 it is revised as the latest version known as "Management regulations of Environmental Analysis Laboratories". Also series of basic criteria for the environmental analysis laboratory's QA system has been formulated according to ISO/IEC 17025 international standards. The accredited laboratories contribute to the environmental analysis of Taiwan to a great extent. There are currently 99 accredited laboratories with more than 3,200 employees that account for more than 90% of lab employees in Taiwan. In terms of the number of analyzed samples, accredited laboratories also account for more than 85 % of all tested samples. Figure 1 shows the chronology of the number of environmental analysis laboratories accredited by the EAL. Figure 2 shows the current category statistics of accredited environmental analysis laboratories, including 48 soil laboratories. Fig. 1. Chronology hronology of the number of environmental analysis laboratories labor accredited by the EAL Fig. 2. Current urrent category statistics of accredited environmental analysis laborator laboratories ESSENTIAL QA/QC REQUIREMENTS IN SOIL ANALYSIS Quality control implies the implementation of monitoring and testing process under regulated criteria. Quality control for the soil testing will follow each specific testing method and eight QA/QC guidelines formulated by EAL. QA/QC guidelines are listed as shown in Table 4. EAL’s Laboratory Quality Manual abides to relevant standard methods, QA/QC guidelines and ISO /IEC 17025. We will brief some of the essential QA/QC requirements in the EAL laboratory regarding soil analysis as following: Preparation of calibration curves When an analytic instrument is used, standard solutions or standard gases in proper concentrations shall be introduced into that instrument to produce response signals before the analysis of real samples. Plot of the response signals vs. the amounts or concentrations of the standards is the so called calibration curve. In general, the concentration of the determined in a sample needs to be adjusted into the concentration range of the calibration curve to be accurately analyzed. Linear calibration curve pass the origin of coordinates: Calculate calibration factors (CF) or response factors (RF), and their mean value and relative standard deviation for various concentrations of the standards used to establish the calibration scope according to the test method. Unless otherwise required of specified standard method, it is assumed that a linear relationship will apply for the calibration curve and the curve will pass through the origin of coordinates if the relative standard deviation of CFs or RFs is not greater than 20%. Average of CFs or RFs will be used to determine the concentration of the sample. Linear calibration curve does not pass the origin of coordinates: Linear regression is made between the instrument response signals and relative concentrations of the standard to establish a calibration curve with a slope and intercept. Linear correlation coefficient for the linear regression is calculated accordingly. The calibration curve must be redone or tried to be set up as a calibration curve in separated concentration range if the linear correlation coefficient is beyond acceptable criteria. Calibration check After the completion of the calibration curve, one needs to check another standard sample solution with a concentration near the mid-range supplied from different sources other than that used to prepare the calibration curve to verify the applicability of the calibration curve. The concentration of that standard obtained from the calibration curve is compared with the known preparation concentration, and a relative error is then calculated out. The appropriateness of a calibration curve is decided according to the following principles or the requirements of individual test method: Unless otherwise requirement of specified standard method, the laboratory can use this calibration curve for testing if the relative error of electrode method falls within ± 15%;colorimetric (spectrophotometric) method, ±10%;ion chromatography, ±15%;flame & graphite atomic absorption spectrometric methods, ± 10%;cold vapor & hydrogen arsenite atomic absorption spectrometric methods, ±20%;inductively couple plasma atomic emission spectrometric method (ICP/AES)& Inductively coupled plasma - mass spectrometry method (ICP/MS), ±10%;gas and liquid chromatographic methods, ±15% and GC/MS methods, ±20%. When the relative error is outside acceptable limits, test that verification standard again. If the retest result is still out of acceptable limits, the standards used for establishing calibration curve shall be prepared again for the establishment of another new one. Determination of instrument detection limit and method detection limit For any newly purchased instrument, it is advisable to request the supplier to determine the instrument detection limit, and relevant records kept on file for future reference. If practicable, it is recommended to review the method detection limits (MDL) every year for those test methods frequently used, and determine their MDLs again every two years at least. Blank sample analysis A field blank sample, trip blank sample or equipment blank sample is analyzed to confirm that the samples are not contaminated during collection and transportation to the laboratory. A method blank sample is prepared in the laboratory and analyzed to confirm the absence of contamination of the samples in the course of testing. Results of blank sample analyses should not be over two times the MDL values, 5﹪of the regulatory compliance limits or the assigned values of test methods, but the responses of blank samples of microbiological testing should be non-detectable. Duplicate sample analysis If applicable, duplicate sample analysis shall be carried out to assess the testing precision. Samples for duplicate analysis should be quantitatively detectable. If the sample is not quantitatively detectable, spike samples or QC check samples may be doubly tested. The results of duplicate sample analysis will define the precision in terms of the Relative Percent Difference (RPD). The RPD should be plotted on control charts for the precision control. Spike sample analysis A spike sample analysis is used to check whether the matrix interference is within the acceptable range or not. Upon carrying out a spike sample analysis, the testing operator shall spike the sample in small volume of standard with high concentration to avoid a significant change to the original sample matrix and may determine the amount spiked in a test sample based on the following principles: 1. To spike the amount about one to five times the concentration of the target analyte in that sample. 2. If the concentration of the target analyte in the sample is completely unpredictable, the spiked amount could be one to five times the background concentration of the target analyte in the field. 3. For the samples taken from pollution areas, the spike amount could be the regulatory compliance limit, half the regulatory compliance limit or that close to the midpoint concentration of calibration curve of the target analyte. The test result of the spike sample versus the theoretical spike recovery is used to calculate the accuracy as defined by the percent recovery. With the above spiking principles, the factor of improper spike, which might affect the recovery, should be taken into account in the evaluation of recovery. Quality control check sample analysis If applicable, each test division shall either purchase or prepare quality control check samples of proper concentration and carry out their analyses together with each batch of samples to evaluating the accuracy of testing. Except required otherwise in the test method, a quality control check sample is generally prepared with the concentration of about the mid-point of the calibration curve. Assessment of control chart Enter test results of quality control samples into LIMS, laboratory information management system, of the EAL for automated plotting of the results into the annually updated various control charts. Properly planned actions shall be taken to prevent incorrect results from being reported in case any of the following events is observed: 1. When the test result of any quality control sample falls beyond lower or upper control limit, the testing for that item shall be immediately suspended until this event is investigated and corrective actions are taken as specified. After finishing the corrective actions, the identity number of the corrective action record sheet would be better recorded on the analyst worksheet to facilitate tracing of the event. 2. When the test results of quality control samples fall between the warning and control limits at the same side of the control chart for two consecutive times, the testing for that item shall be immediately suspended until this event is investigated and corrective actions are taken as specified. 3. When the test results of quality control samples ascend or descend continuously on the control chart for six consecutive times excluding the turning point, the latest batch of test samples should be reanalyzed. If the test result of the quality control sample in the batch reanalyzed falls less than one standard deviation from the mean of the control chart, or changes the direction of ascension or descent, the testing for that item can be continued to carry out. Otherwise, the testing for that item shall be immediately suspended until this event is investigated and corrective actions are taken as specified. 4. When the test results of quality control check samples fall at the same side of the control chart for seven consecutive times, the testing for that item shall be immediately suspended until this event is investigated and corrective actions are taken as specified. In the presence of a certain correlation among different test items of the same sample, the rationality between test results shall be assessed and analyzed for effectively monitoring on testing data quality. Performance audit Performance audit is to evaluate and ensure the testing competence of a laboratory by the way of analyzing internal performance evaluation samples or participating in external inter-laboratory proficiency tests. Whenever applicable, each test division shall periodically participate in overseas proficiency test programs. In case of any testing abnormality identified in the performance audit, the chief or signatory of the relevant division shall initiate a corrective actions record sheet and direct relevant persons to carry out corrective actions. If required, the testing operator may analyze another internal PE Sample for verification as soon as the corrective actions are completed. Internal audit Internal audit relates to checking and assessing testing related activities of each division according to Quality Manual, standard operating procedures and other quality assurance related document to confirm its continuous compliance with the requirements of quality management system. An audit of whole management system shall be carried out at least once each year. Each year, the Executive Officer of internal audit group shall draft the annual internal audit program, which states purpose, auditors, scope, and procedures of audit and preservation of audit records, to be reviewed and resolved in Quality Assurance Team meeting for sequent implementation. Auditors selected from our institute are responsible for carrying out annual internal audits. External specialists could also be invited to help audit, if necessary. The chief(or signatory), Division Quality Assurance officer and relevant personnel of the division audited should be present during the audit. Annual internal audits shall be carried out according to Internal Audit Record Sheet. Auditors are responsible for recording defects, deviations and other findings, and preparing the Internal Audit Report. The chief, signatories or QA officer of the division audited shall complete Internal Audit Registration Record and produce Corrective Action Record to monitor and follow up the corrective action taken for those defects or deviations immediately or within a specified time. The defects or deviations of the division audited that might affect the correctness of the testing results shall be further notified to the applicants. Major essential QA/QC requirements of the EAL laboratory in soil analysis is shown as Table 5. Table4. Eight QA/QC guidelines formulated by the EAL QA/QC guidelines General Rules of Quality Control Guidelines for Environmental Analysis Guidelines of Sampling and Preservation Process for Environmental samples Guidelines of Calibration Curve Preparation and Checkup for Environmental Analysis Quality Control Implementation Guidelines for Environmental Analysis Guidelines of Quality Control Chart Establishment for Environmental Guideline number NIEA-PA101 NIEA-PA102 NIEA-PA103 NIEA-PA104 NIEA-PA105 QA/QC guidelines Guideline number Analysis Guidelines of Laboratory Ware Cleaning and Calibration for Environmental Analysis Guidelines of Method Detection Limit Establishment for Environmental Analysis Guidelines of Instrument Calibration and Maintenance for Environmental Analysis NIEA-PA106 NIEA-PA107 NIEA-PA108 Table5. Major essential QA/QC requirements of the EAL laboratory in soil analysis QA/QC requirement Analytical method Preparation of Calibration Curves Calibration check Blank sample Duplicate sample QC Check Sample Spike sample GC method GC/MS method ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ AA method ○ ○ ○ ○ ○ ○ ICP-AES method HRGC/HRMS method ○ ○ ○ ○ ○ ○ ○ ○ ○ × ○ × Note: "○" is necessary, and "×" i is not necessary. CONCLUSION "Soil and Groundwater Pollution Remediation Act" was officially announced on Feb 2, 2000 in Taiwan. EPA Taiwan established the Soil and Groundwater Pollution Remediation Fund Management Board in November 2001 in accordance with the Act. Based on Article 6 of this Act, soil pollution control standards are issued in 2001. According to Article 10 of the Act, methods and quality control guidelines for soil contaminants analysis are determined by EAL. EAL is also responsible for the accreditation of environmental analysis laboratories. Based on these statutory responsibilities, EAL has issued 91 soil standard testing methods, 8 relevant quality assurance guidelines and granted accreditation to 48 soil environmental testing laboratories following ISO /IEC 17025 guideline. Laboratory quality manual of EAL also strictly follow these standard methods, QA/QC guidelines and ISO /IEC 17025. We have made a briefing on some essential QA/QC requirements relevant in soil analysis. We wish to share and exchange our experiences with all experts of agricultural soil analysis laboratories participating in this seminar. EAL is the authorized department in environmental analysis for the central Government in Taiwan and also a NATA-accredited laboratory. It will review its management system simultaneously with international accreditation standards. We also update all the NATA Accreditation Requirements, internal management practical need and keep make sure to fulfill relevant environmental regulations. We have completed the revision and issuing for related contents in our latest version of Quality Manual. All staff members in EAL, including the management level and every personnel related to analytical works will use the latest version of Quality Manual to carry out all testing works and follow relevant QA/QC requirements. We will do all effort to assure and promote data quality and follow the latest version of ISO/IEC 17025 international standard and meet government environmental regulation. REFERENCES http://www.niea.gov.tw https://sgw.epa.gov.tw/public/ Taiwan EPA, Soil and Groundwater Pollution Remediation Act, 2010. Taiwan EPA, Soil pollution control standards in Taiwan, 2011 Taiwan EPA, Management regulations of Environmental Analysis Laboratories, 2015 Taiwan EPA, Aqua regia digestion method (NIEA S321.64B), 2015. Taiwan EPA, Microwave assisted aqua regia digestion method (NIEA S301.60B), 2015. Taiwan EPA, Flame atomic absorption spectrophotometry method (NIEA M111.01C), 2012. Taiwan EPA, Inductively coupled plasma-atomic emission spectrometry method (NIEA M104.02C), 2013. Taiwan EPA, Inductively coupled plasma - mass spectrometry method (NIEA M105.01B), 2013. Taiwan EPA, hydrogen arsenite atomic absorption spectrometric method (NIEA S310.64B), 2012. Taiwan EPA, Mercury by thermal decomposition, amalgamation, and atomic absorption spectrophotometry Method (NIEA M318.01C), 2012 Taiwan EPA, Mercury by cold vapor atomic absorption spectrophotometry method (NIEA M317.03B), 2016. Taiwan EPA, Closed-system purge-and-trap and extraction for volatile organics method (NIEA M155.01C), 2013. Taiwan EPA, Volatile organic compounds using equilibrium headspace analysis method (NIEA M157.00C), 2016. Taiwan EPA, Volatile organic compounds by gas chromatography/mass spectrometry (GC/MS) method (NIEA M711.04C), 2018. Taiwan EPA, Ultrasonic extraction method (NIEA M167.01C), 2012. Taiwan EPA, Soxhlet extraction method (NIEA M165.01C), 2016. Taiwan EPA, Semivolatile organic compounds by gas chromatography/mass spectrometry (GC/MS): Method (NIEA M731.02C), 2017. Taiwan EPA, Total petroleum hydrocarbons by gas chromatography/flame ionization detector (GC/FID) method (NIEA S703.62B), 2013. Taiwan EPA, Silica gel cleanup method (NIEA M183.01C), 2012. Taiwan EPA, Organochlorine pesticides by gas chromatography method (NIEA M618.05C), 2016. Taiwan EPA, Polychlorinated dibenzo-p-dioxins and dibenzofurans by high-resolution gas chromatography/high-resolution mass spectrometry (HRGC/HRMS) method (NIEA M801.13B), 2013. Taiwan EPA, Polychlorinated dibenzo-p-dioxins and dibenzofurans by gas chromatography/tandam mass spectrometry (GC-MS/MS) method (NIEA M805.00B), 2017. Taiwan EPA, Polychlorinated biphenyls (PCBs) by gas chromatography method (NIEA M619.04C), 2017. Taiwan EPA, General Rules of Quality Control Guidelines for Environmental Analysis -NIEA-PA101, 2004 Taiwan EPA, Guidelines of Sampling and Preservation Process for Environmental samples-NIEA-PA102, 2004 Taiwan EPA, Guidelines of Calibration Curve Preparation and Checkup for Environmental Analysis -NIEA-PA103, 2004 Taiwan EPA, Quality Control Implementation Guidelines for Environmental Analysis -NIEA-PA104, 2004 Taiwan EPA, Guidelines of Quality Control Chart Establishment for Environmental Analysis -NIEA-PA105, 2004 Taiwan EPA, Guidelines of laboratory ware cleaning and calibration for Environmental Analysis -NIEA-PA106, 2004 Taiwan EPA, Guidelines of method detection limit Establishment for Environmental Analysis -NIEA-PA107, 2004 Taiwan EPA, Guidelines of instrument calibration and maintenance for Environmental Analysis -NIEA-PA108, 2004 Taiwan EPA, Soil and groundwater analysis and QA/QC, soil pollution assessment and investigation personnel training textbook, Volume 13, 2017 EAL, Taiwan EPA, Laboratory quality manual, 2017
0
You can add this document to your study collection(s)
Sign in Available only to authorized usersYou can add this document to your saved list
Sign in Available only to authorized users(For complaints, use another form )