CELL-DYN® 3000/3500/3700/3200 TROUBLESHOOTING GUIDE Abbott Diagnostics A Division of Abbott Laboratories Abbott Park, IL 60064 ii CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 REVISION STATUS Revision History Pages Revised and Added 67483-101 6/2002 First Release CELL-DYN is a registered trademark of Abbott Laboratories. This guide was developed and produced by TLO Learning Services in Irving, TX. Copyright 2002, Abbott Laboratories CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 iii Notes CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 iv CELL-DYN® 3000/3500/3700/3200 TROUBLESHOOTING GUIDE CONTENTS RBC Diluent Syringe Overpressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 HGB Lyse Empty (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 WBC Lyse Empty (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Lyse Empty (CELL-DYN 3500/3700) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Diluent/Sheath Empty (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Reagent Empty - Diluent, Sheath, HGB Lyse (CELL-DYN 3000) . . . . . . . . . . . . . . . . . . . . . . . . 9 Reagent Empty - Diluent, Sheath, Detergent (CELL-DYN 3500/3700) . . . . . . . . . . . . . . . . . . . 11 Shear Valve Position Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Vacuum Accumulator Wet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Clog Error (RBC or WIC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Flow Error (RBC or WIC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 WOC / WBC Flow Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Elevated PLT Backgrounds / LRI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Controls Out of Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Controls Out of Range - WBC / WOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Controls Out of Range - NOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Controls Out of Range - WIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Controls Out of Range - RBC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Controls Out of Range - HGB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Controls Out of Range - MCV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 1 Controls Out of Range - PLT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 WIC / WOC Mismatch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 H & H Mismatch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Incomplete Aspiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Increased RRBC Flags . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 Background Data Unacceptable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Data Not Transmitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Reticulocyte Troubleshooting for CELL-DYN 3200, 3500 and 3700 . . . . . . . . . . . . . . . . . . . . 55 WOC Flow Cell Cleaning Procedure (CELL-DYN 3000/3500/3700) . . . . . . . . . . . . . . . . . . . . 59 WOC Flow Cell Cleaning Procedure (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 RBC Diluent Syringe Overpressure RBC DILUENT SYRINGE OVERPRESSURE RBC Diluent RBC Diluent Syringe Syringe Overpressure Overpressure Check center section of shear valve Is notch in correct position? No Remove and reinstall Shear Valve & release pressure from RBC Diluent Syringe (See Note) Run backgrounds Run backgrounds Does error recur? Yes Check RBC dilution pathway for pinched/ crimped tubing No Problem resolved Yes Pinched/ crimped tubing seen? Refer to HS or dispatch service Yes Remove crimps/ massage tubing Go to Diagnostics, More, Initialization Run backgrounds No Does error recur? NOTE: Remove RBC diluent syringe from bracket (do not disconnect tubing) y y y Problem resolved Yes Pull down slightly on plunger Reinitialize analyzer by going to: - Main - Diagnostics - More - Initialization If cannot initialize, cycle power and reinitialize Refer to HS or dispatch service No Remove, clean, and reinstall shear valve & release pressure from RBC diluent syringe (See Note) Run backgrounds Does error recur? No Problem resolved Yes Yes Does error recur? Clean/replace RBC Diluent Syringe No Problem resolved CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 3 HGB Lyse Empty (CELL-DYN 3200) HGB LYSE EMPTY (CELL-DYN 3200) HGB Lyse Empty HGB Lyse3200) Empty (CELL-DYN (CELL-DYN 3200) Press CLEAR FAULT and observe HGB Lyse Syringe Is syringe moving properly? No Clean or replace syringe Does error recur? No Problem resolved Yes Yes Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Is syringe filled properly? Are reagent lines not crimped? Dispatch Service No Disable analyzer, exercise valve 24 using valve puller Does error recur? Massage and reseat tubing in valve 28 Yes Yes Refer to HS or dispatch service Cycle power Dispatch Service 4 Yes Does error recur? No Problem resolved CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 WBC Lyse Empty (CELL-DYN 3200) WBC LYSE EMPTY (CELL-DYN 3200) WBC Lyse Empty WBC Lyse3200) Empty (CELL-DYN (CELL-DYN 3200) Press CLEAR FAULT and observe WBC Lyse Syringe Is syringe moving properly? No Clean or replace syringe Does error recur? Problem resolved Yes Yes Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Is syringe filled properly? Are reagent lines not crimped? No Dispatch Service Disable analyzer, and exercise valve 25 using valve puller No Massage and reseat tubing in valve (normally closed) 23 Yes Refer to HS or dispatch service Does error recur? No Problem resolved No Problem resolved Yes Cycle power Does error recur? Yes Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 5 Lyse Empty (CELL-DYN 3500/3700) LYSE EMPTY (CELL-DYN 3500/3700) L y s e E m p ty ( C E L L - D LYyNs e3 5E0m0 p/3t y7 0 0 ) (C E L L -D Y N 3 5 0 0 /3 7 0 0 ) C h e c k r e a g e n t c u b e a n d in le t lin e s : - c u b e e m p ty ? - r e a g e n t c o r r e c t? - in le t lin e r e a c h e s to b o tto m ? - v e n t h o le in c a p is o p e n ? - fo a m s e a l r e m o v e d fr o m u n d e r c a p ? - lin e s n o t c r im p e d ? - p o r ts n o t p lu g g e d ? - c h e c k flu id le v e ls in s y r in g e P r im e ly s e t o w a s t e c h a m b e r 2 b y r e m o v in g t u b in g f r o m V 3 6 (N C V ) P re s s C L E A R F A U L T F lo s s tu b in g in to N C V 3 6 E x e r c is e s o le n o id s 42, 47 D o e s e rro r re c u r? Yes No D o e s e rro r re c u r? No P r o b le m r e s o lv e d Yes C h e c k W IC /H G B L y s e S y r in g e : - s e a te d a n d m o v in g c o r r e c t ly ? - tu b in g c r im p e d ? C le a n /r e s e a t s y r in g e a s n e e d e d a n d r e m o v e c r im p s D o e s e rro r re c u r? No P r o b le m r e s o lv e d Yes C h e c k ly s e in le t lin e fo r c r im p s o r p lu g s R e m o v e ly s e in le t tu b in g & f lu s h lin e w ith w a rm D I w a te r P r o b le m r e s o lv e d No D o e s e rro r re c u r? Yes No C y c le P o w e r D o e s e rro r re c u r? Yes R e fe r to H S o r d is p a tc h s e r v ic e 6 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Diluent/Sheath Empty (CELL-DYN 3200) DILUENT/SHEATH EMPTY (CELL-DYN 3200) Diluent/Sheath Empty Diluent/Sheath Empty (CELL-DYN 3200) (CELL-DYN 3200) Press CLEAR FAULT and open front left door Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Run background Problem resolved No Does error recur? Yes Press CLEAR FAULT, run background, and observe diluent and sheath reservoirs Does fluid reach the sensor electrodes? No Check reservoirs for cracks / leaks Yes Refer to HS or dispatch service Cracks / leaks seen? No Check sensor contacts at top of reservoirs are seated properly. Reseat if indicated. Run background count. Yes Dispatch service Does error recur? No Problem resolved Yes Page 1 of 2 Go to A CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 7 Diluent/Sheath Empty (CELL-DYN 3200) DILUENT/SHEATH EMPTY (CELL-DYN 3200 ) A Check for salt buildup at reservoir connectors and clean if present Problem resolved No Does error recur? Yes - Verify diluent/sheath plumbing is free of restrictions - Exercise valves 61, 62, 63, and 64 - Are reservoirs cracked and leaking? - Clean sensor contacts (electrodes and connector at top of reservoir) and contacts are touching Run background Problem resolved No Does error recur? Yes Cycle power Problem resolved No Does error recur? Yes Refer to HS or dispatch service Page 2 of 2 8 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Reagent Empty - Diluent, Sheath, HGB Lyse (CELL-DYN 3000) REAGENT EMPTY - DILUENT, SHEATH, HGB LYSE (CELL-DYN 3000) CELL-DYN 3000 CELL-DYN 3000 Reagent Empty Reagent Diluent Empty Diluent Sheath HGBSheath Lyse HGB Lyse Clear fault Press RUN and observe reagent reservoirs Does fluid reach sensor contacts? No Yes Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Clean sensor contacts Ensure sensor is seated correctly Run background Does error recur? Problem resolved No Does error recur? No Problem resolved Yes No Yes Does error recur? Yes Refer to HS or dispatch service Check reservoir for cracks Cycle power Floss NCV tubing (V6-1 (Diluent), V6-2 (Sheath), V6-3 (HGB Lyse)) for affected reservoir or replace check valve No Cracks seen? Yes Problem resolved No Does error recur? Refer to HS or Dispatch Service Yes Page 1 of 2 Go to A CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 9 Reagent Empty - Diluent, Sheath, HGB Lyse (CELL-DYN 3000) REAGENT EMPTY - DILUENT, SHEATH, HGB LYSE (CELL-DYN 3000) A Exercise vacuum and pressure solenoids for affected reservoir Diluent pressure and vacuum - V5-1, 5-2 Sheath pressure and vacuum - V5-3, 5-4 HGB Lyse pressure and vacuum - V5-5, 5-6 Problem resolved No Does error recur? Yes Cycle power Problem resolved No Does error recur? Yes Page 2 of 2 10 Refer to HS or Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Reagent Empty - Diluent, Sheath, Detergent (CELL-DYN 3500/3700) REAGENT EMPTY - DILUENT, SHEATH, DETERGENT (CELL-DYN 3500/3700) CELL-DYN 3500/3700 CELL-DYN 3500/3700 Reagent Empty Reagent Diluent Empty Diluent Sheath Sheath Detergent Detergent Clear fault Press RUN and observe reagent reservoirs Does fluid reach sensor contacts? No Yes Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Clean sensor contacts Ensure sensor is seated correctly Does error recur? Problem resolved No Does error recur? No Problem resolved Yes No Yes Does error recur? Check reservoir for cracks Cycle power Yes Floss NCV tubing for affected reservoir or replace check valve Refer to HS or dispatch service No Cracks seen? Yes Problem resolved No Does error recur? Refer to HS or Dispatch Service Yes Page 1 of 2 Go to A CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 11 Reagent Empty - Diluent, Sheath, Detergent (CELL-DYN 3500/3700) REAGENT EMPTY - DILUENT, SHEATH, DETERGENT (CELL-DYN 3500/3700) A Exercise vacuum and pressure solenoids for affected reservoir V5-1 Diluent pressure V5-2 Diluent vacuum V5-3 Sheath pressure V5-4 Sheath vacuum V5-5 Detergent pressure V5-6 Detergent vacuum Problem resolved No Does error recur? Yes Cycle power Problem resolved No Does error recur? Yes Page 2 of 2 12 Refer to HS or Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Shear Valve Position Fault SHEAR VALVE POSITION FAULT S h e a r V a lv e h ena rF V a lt lv e P o sSitio au P o s itio n F a u lt D id e rro r o c c u r a fte r c le a n in g s h e a r v a lv e ? C le a n a n d in s p e c t s h e a r v a lv e No Yes Is s h e a r v a lv e d a m a g e d /p in s b ro k e n ? Yes R e fe r to H S o r D is p a tc h s e rv ic e No R e m o v e a n d re in s ta ll s h e a r v a lv e c o rre c tly (C E L L -D Y N 3 0 0 0 : u s e lu b ric a n t) R e in s ta ll s h e a r v a lv e c o rre c tly R e in itia liz e s y s te m D o e s e rro r re c u r? No P ro b le m re s o lv e d Yes C y c le p o w e r D o e s e rro r re c u r? No P ro b le m re s o lv e d Yes R e fe r to H S o r d is p a tc h s e rv ic e CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 13 Vacuum Accumulator Wet VACUUM ACCUMULATOR WET Vacuum Vacuum Accumulator W et Accumulator W et Observe waste line for obstructions or correct draining Is waste line draining correctly? No Remove obstruction/reseat tubing Yes Drain accumulators CELL-DYN 3000: Follow Operator's Manual for manual procedure CELL-DYN 3500/3700/3200: through DIAGNOSTICS menu (go to Diagnostics, More, Drain accumulator) Reinitialize system Does error recur? No Problem resolved No Problem resolved Yes Cycle power Does error recur? Yes Refer to HS or dispatch service 14 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Clog Error (RBC or WIC) CLOG ERROR (RBC OR WIC) C log E rror (R B C or W IC ) C log E rror (R B C or W IC ) R un background P roblem resolved No D oes error recur? P roblem resolved Y es C heck reagent cube and inlet lines: - cube em pty? - reagent correct? - inlet line reaches to bottom ? - vent hole in cap is open? - foam seal rem oved from under cap? - lines not crim ped? - ports not plugged? - check fluid level in syringe No D oes error recur? Y es P roblem resolved No R B C C log: - clean/replace 10 m L syringe - clean air filter - run autoclean D oes error recur? V erify aperture plate(s) are correctly installed and check for cracks R eseat and clean per O P S m anual Y es E xercise solenoids around affected transducers and m assage and reseat tubing at solenoids (S ee N ote) NOTE: S olenoids C E LL-D Y N 3000: 3-1, 3-2, 3-3, 3-6, 3-7, 3-8 D oes error recur? C E LL-D Y N 3500/3700: R B C : 3-1, 3-2, 3-3, 3-4, 3-5, 3-7, 3-8 W IC : 8-1, 8-2, 8-3, 8-4, 8-5, 8-7, 8-8 Y es No P roblem resolved G o to A P age 1 of 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 15 Clog Error (RBC or WIC) CLOG ERROR (RBC OR WIC) A Check transducer(s) and metering tubes for cracks or leaks Check quality of meniscus in metering tube No Cracks/leaks seen? Yes Check fitting between top of metering tube and V 37 (RBC) or V 8-7 (W IC) No Good meniscus? Dispatch Service Clean or flush metering tube if needed Yes Problem resolved Page 2 of 2 16 No Does error recur? Problem resolved Yes No Cycle power Does error recur? Yes Refer to HS or dispatch service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Flow Error (RBC or WIC) FLOW ERROR (RBC OR WIC) Flow Error Error (RBCFlow or WIC) (RBC or WIC) Run background Problem resolved No Does error recur? Yes Inspect transducer/vent solenoids and pinch tubing Massage and reseat tubing in solenoids as needed SEE NOTE Problem resolved No Does error recur? Yes Verify aperture plate(s) are correctly installed and not cracked Reseat/replace if indicated Problem resolved No Does error recur? Yes Check transducer(s) and metering tubes for cracks or leaks NOTE: Solenoids CELL-DYN 3000: 3-1, 3-2, 3-3, 3-6, 3-7, 3-8 CELL-DYN 3500/3700: RBC: 3-1, 3-2, 3-3, 3-4, 3-5, 3-7, 3-8 WIC: 8-1, 8-2, 8-3, 8-4, 8-5, 8-7, 8-8 Cracks/leaks seen? No Go to A Yes Page 1 of 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Dispatch Service 17 Flow Error (RBC or WIC) FLOW ERROR (RBC OR WIC) A Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Run background Problem resolved No Does error recur? Yes Note: To clean RBC or WIC metering tube: CELL-DYN 3000: V38 (RBC) CELL-DYN 3500/3200: V38 (RBC), V88 (WIC) Check quality of meniscus in metering tube -- it should look like an upside down U. Check fitting between top of metering tube and V 3-7 (RBC) or V 8-7 (WIC)and exercise valves if indicated Clean metering tubes (See Note) No 1. Disconnect tubing above V38 or V88. 2. Connect 20 mL syringe filled with 20% bleach to tubing. 3. Open V38 or V88 until syringe is half empty. 4. Release solenoid and wait 5 minutes. 5. Open V38 or V88 until syringe is empty and release solenoid. 6. Fill syringe with HGB Lyse reagent and repeat step 5. 7. Run background. Good meniscus? Yes Problem resolved No Does error recur? Yes Run autoclean Does error recur? No Problem resolved No Problem resolved Yes Cycle power Refer to HS or Dispatch Service Yes Does error recur? Page 2 of 2 18 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 WOC / WBC Flow Errors WOC / WBC FLOW ERRORS WOC/WBC Flow WOC/WBC Flow Errors Errors Is error occurring on one sample only? Yes No Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Rerun sample Does error recur? Yes Problem resolved No Does error recur? Yes Verify Results by an alternate method Drain/Fill reagents Does error recur? No Replace reagents No Problem resolved No Clean or replace as needed Yes Verify syringes: - seated correctly - moving correctly - bubble free See NOTE Syringe OK? Yes NOTE: Check the following syringes: CELL-DYN 3000: WBC metering syringe CELL-DYN 3200: Sample injection syringe CELL-DYN 3500/3700: WOC metering sheath syringes Go to A Does error recur? No Problem resolved Yes Page 1 of 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 19 WOC / WBC Flow Errors WOC / WBC FLOW ERRORS A Check transfer peristaltic pump tubing for wear or holes Replace as needed SEE NOTE Does error recur? No Problem resolved No Check for pinched tubing Exercise bubble mix solenoid 2-4 Yes Verify WOC/WBC mixing chamber fills correctly, bubble mix is good (Swirl mix for CELL-DYN 3200) Mixing chamber OK? Yes Dispatch Service No Check channel offsets (<2.0, N/A on CELL-DYN 3200) Run background Offsets in specs? Does error recur? No Problem resolved No Yes Yes Yes Cycle power Clean Shear Valve Does error recur? Yes Does error recur? Cycle power Yes Does error recur? No Problem resolved Dispatch Service No Problem resolved Page 2 of 2 20 NOTE: CELL-DYN 3000: WBC transfer peripump tubing CELL-DYN 3200: Sample transfer peripump tubing CELL-DYN 3500/3700: WOC transfer peripump tubing CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Elevated PLT Backgrounds / LRI ELEVATED PLT BACKGROUNDS / LRI Elevated PLT Elevated/ PLT backgrounds LRI backgrounds / LRI Check that ground wire is attached, front covers on Repeat background Problem resolved No Does problem recur? Yes Run electrical background Go to C No No Is it a CELLDYN 3200? Yes Empty/fill flow cell Is electrical background = 0? Does problem recur? No Yes Yes Check diluent/sheath reagent: - correct list #? - frozen? Replace if needed and prime twice after replacing. Check diluent/sheath reagent: - correct list #? - frozen? Replace if needed and prime twice after replacing Problem resolved No Problem resolved Does problem recur? Yes Run autoclean Problem resolved No Does problem recur? Yes Go to A Page 1 of 4 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 21 Elevated PLT Backgrounds / LRI ELEVATED PLT BACKGROUNDS / LRI A C h e c k D ilu e n t (D ilu e n t/S h e a th s y rin g e fo r C E L L -D Y N 3 2 0 0 ) s y rin g e fo r le a k s / b u b b le s /b u ild u p C le a n o r re p la c e if in d ic a te d R u n b a c k g ro u n d P ro b le m re s o lv e d No D o e s p ro b le m re c u r? Yes C le a n R B C /P L T A p e rtu re P la te and (C E L L -D Y N 3 0 0 0 /3 5 0 0 /3 7 0 0 o n ly ) s h e a r v a lv e a s p re c a u tio n R u n b a c k g ro u n d P ro b le m re s o lv e d No D o e s p ro b le m re c u r? Yes C E L L -D Y N 3 0 0 0 /3 2 0 0 : C h e c k / re p la c e (if in d ic a te d ) d ilu e n t/ s h e a th filte r R u n b a c k g ro u n d P ro b le m re s o lv e d No D o e s p ro b le m re c u r? Yes D ra in /fill b a th s a n d re s e rv o irs (flo w c e ll o n C E L L -D Y N 3 2 0 0 ) P ro b le m re s o lv e d No D o e s p ro b le m re c u r? Yes G o to B P age 2 of 4 22 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Elevated PLT Backgrounds / LRI ELEVATED PLT BACKGROUNDS / LRI B Perform prepare for shipping procedure Run 3-5 backgrounds Problem resolved No Does problem recur? Yes Clean or drain vacuum accumulators Run background Problem resolved No Does problem recur? Yes Clean fan filters Run background NOTE: Transducer cleaning Problem resolved No Does problem recur? Problem resolved No Does problem recur? Yes Clean waste line Run background 1. From Main menu, press: Special Protocols, Empty Xducers 2. Disconnect vent tubing at top of transducer (at top right port of first transducer) 3. Connect 20 mL syringe filled with 20% bleach solution to top of the port 4. Inject bleach solution until the left side of transducer is half filled and reconnect tubing. 5. Allow bleach to soak for 5 minutes. 6. Press: Fill xducers Main menu Run 7. Run 5 backgrounds Yes Check transducer electrodes for corrosion Clean transducer if indicated (to clean transducer, see note:) Does problem recur? Yes Cycle power No Does problem recur? No Problem resolved Yes Refer to HS or Dispatch Service Page 3 of 4 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 23 Elevated PLT Backgrounds / LRI ELEVATED PLT BACKGROUNDS / LRI C Check for power cords resting on diluent or waste lines Remove if found Run background Problem resolved No Does problem recur? Yes Check waste sensor dummy plug (if used) Reseat plug Problem resolved No Does problem recur? Yes Clean fan filters, drain or clean vacuum accumulators, and waste line Run background Problem resolved No Does problem recur? Yes Check environment: - printer/other devices on or near CELL-DYN? (e.g.: centrifuge, air conditioning) - flickering/fluorescent lights? - humidity? - CELL-DYN connected to UPS? Problem resolved Remove potential interferences Problem resolved No Does problem recur? Yes No Does problem recur? Cycle power Yes Page 4 of 4 24 Refer to HS or Dispatch Service Yes CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Controls Out of Range CONTROLS OUT OF RANGE Controls Out of Controls Range Out of Range - Verify QC means/limits - Check storage/handling - Condition of QC when received - Rerun controls Does problem recur? No Problem resolved Check reagent cube and inlet lines: - cube em pty? - reagent correct? - inlet line reaches to bottom ? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Yes Are backgrounds within specs? Troubleshoot backgrounds No Yes No Does problem occur in one mode only? Run Fresh Controls Yes Does problem recur? No Problem resolved Yes Does problem recur? Troubleshoot aspiration system of affected system Yes No Problem resolved No Go to Incomplete Aspiration Tree No Problem resolved Check sample aspiration through Shear Valve Does problem recur? No Yes Are sam ples aspirated correctly? Run precision Yes Troubleshooting im precision No Precision within specs? Yes Run autoclean Yes Verify mode to mode calibration Recalibrate if needed Does problem recur? Yes Problem resolved No Does problem recur? Go to OPEN / CLOSED m ode Page 1 of 10 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 25 Controls Out of Range CONTROLS OUT OF RANGE Problem occuring in OPEN or CLOSED mode Issue resolved OPEN / CLOSED Were Controls rerun? No No Controls still out of range? Try a different box of controls Yes Controls still out of range? No Yes Yes Issue resolved Check affected parameters Yes No Were controls mixed by hand, warmed to room temperature for 15 minutes, and stored inside refrigerator shelf? No Are multiple parameters out of range? Redo controls mixed by hand No Yes Follow flow chart for affected parameter Clean shear valve, clean Y-fitting (for CELL-DYN 3000/3500/3700) or Y-valve (for CELL-DYN 3200) and rerun Controls Controls within range? WOC/ WBC NOC WIC RBC HGB MCV PLT Go to A Go to B Go to C Go to D Go to E Go to F Go to G Yes Problem resolved No Does problem recur? Yes Issue resolved Yes Check syringes: - seated/moving properly? - filling? Clean/replace as needed Problem resolved No Does problem recur? Page 2 of 10 26 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Controls Out of Range - WBC / WOC CONTROLS OUT OF RANGE - WBC / WOC A Run precision Rerun precision Verify Sample Injection Syringe and/or WOC Metering Syringe are not leaking, have no bubbles, and are moving up/ down properly (visually inspect) No - change Transfer Pump Tubing Is precision within range? Precision within specs? Yes Empty/fill WOC flow cell and rerun controls No Issue resolved Yes Yes Problem resolved No Issue resolved Does problem recur? Is placement of neutrophil population correct on 0/10 scattergram? Yes No Clean WOC flow cell or perform clean for ship and rerun controls Controls in range? Run normal patient sample No Replace WOC peripump tubing and rerun controls Yes Check channel offsets (CELL-DYN 3000/3500/3700 only) Dispatch service No Are channel offsets < 2.00? Yes Verify correct operation of syringes (see NOTE) Clean or replace as needed NOTE: Check the following syringes: CELL-DYN 3000: WBC metering syringe & WOC sheath syringe CELL-DYN 3200: Sample injection syringe CELL-DYN 3500/3700: WOC metering syringe & WOC sheath syringe Does problem recur? No Problem resolved Yes Page 3 of 10 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Go to AA 27 Controls Out of Range - WBC / WOC CONTROLS OUT OF RANGE - WBC / WOC AA Check WBC/WOC mixing chamber for proper draining, filling & bubble mix Exercise mix chamber solenoids Dispatch service No Is mixing chamber draining/filling correctly? Yes Verify WBC/WOC calibration by performing a calibration verification and calibrate if indicated Run controls Problem resolved No Is the WOC within ranger? Yes Dispatch service Page 4 of 10 28 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Controls Out of Range - NOC (CELL-DYN 3200 Only) CONTROLS OUT OF RANGE - NOC (CELL-DYN 3200 ONLY) B Run precision Rerun precision Verify sample injection syringe or WOC metering syringe is not leaking, has no bubbles, and is moving up/down properly (visually inspect) No - change Transfer Pump Tubing Is precision within range? Is precision within specs? Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Yes Are HGB results affected? No Yes Yes Issue resolved Clean shear valve and HGB flow cell No Check solenoid 41 Massage and reseat tubing Does problem recur? No Problem resolved No Problem resolved Yes Problem resolved Problem resolved Does problem recur? No Yes Clean/replace as needed Check HGB Lyse Syringe: - seated/moving properly? - filling? Clean or replace as needed Does problem recur? No Does problem recur? Yes Check staging pathway from shear valve to HGB flow cell for obstructions Problem resolved Check sample injection syringe No Does problem recur? Yes Verify NOC calibration by performing a calibration verification and calibrate if indicated Rerun controls Does problem recur? No Problem resolved Yes Refer to HS or dispatch service Yes Page 5 of 10 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Refer to HS or dispatch service 29 Controls Out of Range - WIC CONTROLS OUT OF RANGE - WIC C Refer to HS or dispatch service Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Run Precision No Is precision within range? Troubleshoot imprecision by cleaning WIC aperture plate Redo precision No Is precision within specs? Yes Rerun controls Yes Problem resolved Yes Is WIC within range? No Verify transducer operation by emptying, filling, draining & bubble mix Exercise transducer solenoids, massage and reseat tubing Refer to HS or dispatch service No See chart of solenoids to exercise Solenoids 8-1, 8-2, 8-3, 8-4, 8-5, and 8-7 Is transducer operating correctly? Yes Check WIC/HGB Lyse and Diluent syringes Clean/replace as needed Problem resolved No Does problem recur? Yes Verify WIC calibration by performing a calibration verification Calibrate if indicated Rerun controls Problem resolved No Does problem recur? Yes Refer to HS or dispatch service Page 6 of 10 30 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Controls Out of Range - RBC CONTROLS OUT OF RANGE - RBC D Run precision Troubleshoot Imprecision Clean shear valve (ensure inserted correctly) Replace sample peripump tubing (except CELL-DYN 3200) Is precision within specs? No Yes Does problem recur? Clean RBC aperture plate (except CELL-DYN 3200) No Problem resolved Yes Problem resolved Massage tubing through solenoid 5-4 and verify RBC chamber is draining, filling, and swirl mixing properly for CELL-DYN 3200 Does problem recur? No Yes Solenoids 3-1, 3-2, 3-3, 3-4, 3-5, and 3-7 Verify transducer operation Exercise transducer solenoids, massage and reseat tubing (except CELL-DYN 3200) Refer to HS or dispatch service Is transducer operating correctly? Does problem recur? No Problem resolved No Refer to HS or dispatch service No Problem resolved Yes No CELL-DYN 3200: check vacuum #2 (3.5" open, 5.0" closed) Yes CELL-DYN 3200 Verify RBC mixing chamber for proper fill/drain; swirl mix Vac #2 within specs? Check Diluent syringe Clean/replace as needed Problem resolved No Yes Verify RBC calibration by performing calibration verification using calibrator Calibrate if indicated Does problem recur? Refer to HS or dispatch service Yes Does problem recur? Page 7 of 10 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 31 Controls Out of Range - HGB CONTROLS OUT OF RANGE - HGB E Issue resolved Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Run precision Yes Troubleshoot HGB imprecision by cleaning HGB Flow Cell and rerun precision Is precision within range? No Is precision within specs? Yes Check HGB reference and HGB voltage within specs and document No HGB reference within specs? No Clean HGB flow cell Refer to HS or dispatch service No HGB reference within specs? Yes Problem resolved No HGB Reference >2200 for CELL-DYN 3000/ 3500/3700 >2050 for CELL-DYN 3200 No Yes Refer to HS or dispatch service Check HGB syringes Check: CELL-DYN 3200 Solenoids 2-4 -- 2-8 and massage tubing CELL-DYN 3500/3700 Solenoid 3-6 Clean/replace as needed Yes Rerun controls Does problem recur? Does problem recur? No Problem resolved Yes Verify HGB calibration by performing the calibration verification Calibrate if indicated Refer to HS or dispatch service Yes Does problem recur? No Problem resolved Page 8 of 10 32 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Controls Out of Range - MCV CONTROLS OUT OF RANGE - MCV F Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Run precision Troubleshoot Imprecision No Is precision within specs? Yes Clean RBC Aperture Plate and RBC / PLT Transducer (CELL-DYN 3000/3500/3700 only) CELL-DYN 3200: verify RBC mixing chamber is draining and swirl mixing properly Replace Diluent / Sheath if indicated Problem resolved No Does problem recur? Yes Check RER data (CELL-DYN 3000/3500/3700 only) 25-35% for normal patient Refer to HS or dispatch service No Is RER data within specs? Yes Verify MCV calibration by performing calibration verification Calibrate if indicated Rerun controls Problem resolved No Does problem recur? Yes Refer to HS or dispatch service Page 9 of 10 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 33 Controls Out of Range - PLT CONTROLS OUT OF RANGE - PLT G Run precision Troubleshoot imprecision by cleaning RBC aperture plate or replacing Diluent / Sheath filter (for CELL-DYN 3200) Rerun precision Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Is precision within specs? No CELL-DYN 3200: RBC / PLT mix chamber; swirl Yes Problem resolved Is precision within range? Yes No Refer to HS or dispatch service Rerun controls CELL-DYN 3200: massage tubing through solenoid 5-4 Does problem recur? Does problem recur? Yes Yes Verify transducer operation Exercise transducer solenoids, massage and reseat tubing (CELL-DYN 3000/3500/3700 only) Verify PLT calibration by performing calibration verification No No Is transducer operating correctly? No Problem resolved No Problem resolved Calibrate if indicated Rerun controls Yes Does problem recur? Yes Yes Check Diluent syringe Clean/replace as needed Refer to HS or dispatch service Issue resolved Problem resolved No Does problem recur? Page 10 of 10 34 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 WIC / WOC Mismatch (CELL-DYN 3500/3700 Only) WIC / WOC MISMATCH (CELL-DYN 3500/3700 ONLY) W IC / W OC Mism atch (m ultiple sam ples) Run controls in open m ode Are W IC and W OC correct? No Yes Yes Troubleshoot closed m ode Troubleshoot W OC No No Troubleshoot W IC Go to D Go to A Run precision Troubleshoot im precision Is W IC correct? Precision within specs? Yes Check closed m ode aspiration See Incom plete Aspiration tree Yes Problem resolved No Does problem recur? Yes Verify m ode to m ode calibration Calibrate if indicated Problem resolved No Does problem recur? Page 1 of 6 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 35 WIC / WOC Mismatch (CELL-DYN 3500/3700 Only) WIC / WOC MISMATCH A Run precision Troubleshoot imprecision No Precision within specs? Problem resolved No Does problem recur? Yes Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Yes Check W OC Sheath and metering syringes: - filled/seated/ moving properly? Clean/replace if needed Problem resolved No Does problem recur? Yes Check sample aspirations/shear valve Go to Incomplete Aspiration tree No Is sample aspirating correctly? Yes Page 2 of 6 36 Go to B CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 WIC / WOC Mismatch (CELL-DYN 3500/3700 Only) WIC / WOC MISMATCH B Check WOC mixing chamber Check solenoids 1-7, 2-3, 2-4 Massage and reseat tubing Does chamber drain/fill and bubble mix properly? No Yes Problem resolved No Does problem recur? Problem resolved Replace WOC peristaltic pump tubing Yes No Does problem recur? Yes Check staging pathway from WOC mixing chamber to flow cell: Check solenoids 2-1, 2-2, 2-3 Massage and reseat tubing Problem resolved No Does problem recur? Yes Massage and reseat tubing in solenoids 2-5, 2-7 Problem resolved No Does problem recur? Go to C Page 3 of 6 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 37 WIC / WOC Mismatch (CELL-DYN 3500/3700 Only) WIC / WOC MISMATCH C Check channel offsets Are channel offsets less than 2.00? No Dispatch service or refer to HS Yes Verify WOC calibration Recalibrate if indicated Problem resolved No Does problem recur? Yes Page 4 of 6 38 Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 WIC / WOC Mismatch (CELL-DYN 3500/3700 Only) WIC / WOC MISMATCH D Run Precision Troubleshoot imprecision No Precision within specs? Problem resolved No Does problem recur? Yes Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Yes Check WIC/HGB diluent and lyse syringes: - filled/seated/moving properly? Clean/replace if needed Problem resolved No Does problem recur? Yes Clean WIC Aperture Plate Problem resolved No Does problem recur? Yes Go to E Page 5 of 6 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 39 WIC / WOC Mismatch (CELL-DYN 3500/3700 Only) WIC / WOC MISMATCH E Check W IC transducer Check solenoids 8-3, 8-4 Massage and reseat tubing Problem resolved No Does problem recur? Problem resolved Is left cham ber of W IC transducer draining/filling properly? No Check tubing between top of transducer and vent accum ulator for crim ps Massage and reseat tubing Yes No Does problem recur? Yes Check sam ple aspiration/shear valve Go to Incom plete Aspiration tree No Is sam ple aspirating correctly? Yes Verify W IC calibration Recalibrate if indicated Problem resolved No Does problem recur? Yes Page 6 of 6 40 Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 H & H Mismatch H & H MISMATCH H & H Mismatch H & H Mismatch Run controls to determine affected parameters Run precision Is precision within specs? No Troubleshoot imprecision Yes Affected parameter MCV RBC HGB Go to A Go to B Go to C Page 1 of 5 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 41 H & H Mismatch H & H MISMATCH C heck reagent cube and inlet lines: - cube em pty? - reagent correct? - inlet line reaches to bottom ? - vent hole in cap is open? - foam seal rem oved from under cap? - lines not crim ped? - ports not plugged? - check fluid level in syringe A P roblem resolved No D oes problem recur? Y es R un autoclean P roblem resolved No D oes problem recur? Y es C lean R B C /P LT aperture plate (C E LL-D Y N 3000/3500/3700 only) P roblem resolved No D oes problem recur? Y es C lean fan filters P roblem resolved No D oes problem recur? Y es P erform C alibration verification C alibrate if indicated P roblem resolved No D oes problem recur? Y es D ispatch S ervice P age 2 of 5 42 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 H & H Mismatch H & H MISMATCH B Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Problem resolved No Does problem recur? Yes Run autoclean Problem resolved No Does problem recur? Yes Check Diluent syringe: - Moving/filling OK? - Bubble free? Clean/replace syringe Problem resolved No Perform Calibration verification Calibrate if indicated Does problem recur? Yes Clean RBC/PLT aperture plate (CELL-DYN 3000/3500/3700 only) Problem resolved No Does problem recur? Yes Does problem recur? No Problem resolved Yes Dispatch Service Page 3 of 5 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 43 H & H Mismatch H & H MISMATCH C Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Problem resolved No Does problem recur? Yes Run autoclean Problem resolved No Does problem recur? Yes NOTE: HGB References: CELL-DYN 3000 1850-2250 CELL-DYN 3500 1850-2250 CELL-DYN 3700 1800-2300 CELL-DYN 3200 1800-2300 Check HGB Reference See Note Clean HGB flow cell No HGB Reference within specs? Yes Refer to HS or Dispatch Service No HGB Reference within specs? Check HGB lyse syringe: - Moving/filling OK? - Bubble free? Yes Yes Problem resolved No Does problem recur? Clean/replace if needed Does problem recur? No Problem resolved Yes Page 4 of 5 44 Go to D CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 H & H Mismatch H & H MISMATCH D Exercise solenoids: CELL-DYN 3000/3500/3700: V 1-3 CELL-DYN 3200: V2-4 Massage/reseat tubing Problem resolved No Does problem recur? Yes CELL-DYN 3500/3700: Exercise WIC transducer drain solenoids V8-2 and V8-3 Massage and reseat tubing Problem resolved No Does problem recur? Yes Perform calibration verification Calibrate if indicated Problem resolved No Does problem recur? Yes Dispatch Service Page 5 of 5 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 45 Incomplete Aspiration INCOMPLETE ASPIRATION Incomplete Incomplete Aspiration Aspiration (multiple samples) (multiple samples) Does error occur in both modes? Go to A No Yes Issue resolved Refer to HS or Dispatch Service CELL-DYN 3000/3500/3700 CELL-DYN 3200 Inspect/change sample peristaltic pump tubing Check vacuum #2 Does error recur? Vac #2 within specs? Yes Yes Check sample sensor Flush Y-valve LED on/off appropriately? Does error recur? No No No Refer to HS or dispatch service No Problem resolved Yes Inspect/clean shear valve and flush Y-fitting with 10% bleach Problem resolved No Does error recur? Yes Yes Refer to HS or Dispatch Service Page 1 of 2 46 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Incomplete Aspiration INCOMPLETE ASPIRATION A Closed Mode only Open Mode only Perform Sam ple path cleaning and include: CELL-DYN 3000/3500/3700 : clean Y-fitting CELL-DYN 3200: clean Y-valve Perform Sam ple path cleaning and include: CELL-DYN 3000/3500/3700 : clean Y-fitting CELL-DYN 3200: clean Y-valve Does error recur? Does error recur? Problem resolved No Check and exercise solenoids: CELL-DYN 3000/3500/3700: 9-5 (open m ode) CELL-DYN 3200: Exercise V1-7 and verify no restrictions between Y-valve and V1-7 Massage/reseat tubing Check/clean sam ple/vent needles No Problem resolved Yes Yes Problem resolved No Does error recur? Does error recur? No Problem resolved No Issue resolved Yes Yes Replace aspiration peripum p tubing (for CELL-DYN 3000/3500/3700 only) Check SL LED Refer to HS or Dispatch Service Sensor LED on/off appropriately? (except CELL-DYN 3200) No Run a Sam ple Yes Exercise CELL-DYN 3000/3500/3700: solenoid 9-6 CELL-DYN 3200: exercise V1-2 and verify no restrictions between Y-valve and V1-2 m assage/ reseat tubing Does error occur? Yes Refer to HS or dispatch service Run another Sam ple Yes Problem resolved No Does error recur? Yes Refer to HS or Dispatch Service Does LED sensor to right of shear valve turns on/off? No Page 2 of 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Refer to HS or dispatch service Yes Does error recur? No Issue resolved 47 Increased RRBC Flags INCREASED RRBC FLAGS Increased RRBC Increased Flags RRBC Flags Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level in syringe Problem resolved No Does problem recur? Yes Clean Shear Valve Problem resolved No Does problem recur? Yes Replace W OC peristaltic pump tubing Problem resolved No Does problem recur? Yes Check W OC sheath and metering syringes (on CELL-DYN 3200, it is the sample injection syringe): - seated and moving properly? - bubble free? Clean/replace if needed Problem resolved No Does problem recur? Yes Page 1 of 3 48 Go to A CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Increased RRBC Flags INCREASED RRBC FLAGS A Check WOC mixing chamber (on CELL-DYN 3500 and 3700 only) or clean it if indicated - see WOC Chamber Cleaning Procedure page 60 (on CELL-DYN 3200: check WBC mixing chamber) For CELL-DYN 3000/3500/3700: Check solenoids 1-7, 2-3, 2-4 and massage and reseat tubing For CELL-DYN 3200: Exercise solenoids 53, 55, 92 and massage and reseat tubing Problem resolved No Does problem recur? No Does chamber drain/fill and bubble mix properly? (on CELL-DYN 3200: does it swirl mix and drain properly?) Yes Check staging pathway from WOC mixing chamber to flow cell: Yes Massage and reseat tubing in staging solenoids Problem resolved No Does problem recur? Yes Check channel offsets (CELL-DYN 3000/ 3500/3700 only) Are channel offsets less than 2.00? Yes Check 0°/10° scattergram Are lymph/neut populations - correctly positioned? - well defined? No Dispatch Service Yes Page 2 of 3 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Go to B 49 Increased RRBC Flags INCREASED RRBC FLAGS B Verify WOC calibration Recalibrate if indicated Problem resolved No Does problem recur? Yes Page 3 of 3 50 Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Background Data Unacceptable BACKGROUND DATA UNACCEPTABLE Background Data Background Data Unacceptable Unacceptable Check that ground wire is attached, front covers on. Repeat background Problem resolved Repeat background OK? Yes No Run electrical background Check for electrical interference* No No Is it a CELLDYN 3200? Is electrical background OK? Yes Empty/fill flow cell Does problem recur? No Problem resolved No Problem resolved Yes Yes Does problem recur? Replace milipore filter Clean Aperture Plates No Yes Problem resolved No Does problem recur/ Does problem recur? Yes Dispatch Service Problem resolved Clean Transducers NOTE: See Elevated PLT Backgrounds/LRI flow chart for Transducer cleaning instructions No Does problem recur? Yes Yes *Examples of electrical interference: - reagent lines near electrical wires - printer on top of CELL-DYN - electrical appliances near CELLDYN - flickering fluorescent lights Go to A Page 1 of 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 51 Background Data Unacceptable BACKGROUND DATA UNACCEPTABLE A Check reagent cube and inlet lines: - cube empty? - reagent correct? - inlet line reaches to bottom? - vent hole in cap is open? - foam seal removed from under cap? - lines not crimped? - ports not plugged? - check fluid level inNo syringe Replace reagent and prime twice Clean vacuum accumulator Clean fan filters Repeat background OK? Repeat background OK? Yes Yes Problem resolved No Cycle power No Perform prepare for ship Problem resolved Yes Repeat background OK? Yes Repeat background OK? No Page 2 of 2 52 Dispatch Service CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Data Not Transmitting DATA NOT TRANSMITTING Data not Data not transmitting transmitting Start Has it worked before? No First time to be hooked-up to an LIS? Yes No Using current rev of interface spec? No Send current interface spec No Problem resolved Is set-up correct? No Correct set-up Does problem recur? No Problem resolved Yes Does problem recur? Yes Yes Check computer set-up (see NOTE) Cable connected to proper port Yes Yes Perform serial loop back test Go to A NOTE: Refer to Operator's Manual for computer setup Page 1 of 2 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 53 Data Not Transmitting DATA NOT TRANSMITTING A Pass? Failed? Yes Yes Dispatch Service Cycle power and reseat LIS cable Does problem recur? No Problem resolved Yes Refer to LIS vendor Page 2 of 2 54 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Reticulocyte Troubleshooting for CELL-DYN 3200, 3500 and 3700 RETICULOCYTE TROUBLESHOOTING FOR CELL-DYN 3200, 3500 AND 3700 Reticulocyte Reticulocyte Troubleshooting for Troubleshooting CELL-DYN 3200, for CELL-DYN 3200, 3500 and 3700 3500 and 3700 Verify that stain has been stored in the dark at room temperature, that Reticulocyte Backgrounds <100 and exact dilution and incubation instructions have been followed. Are calibrated pipettes used? Go to A Both Rerun QC Problems with patients or QC out of range or both? Patients Only QC Only Is it sample specific? No Go to A Yes Issue resolved No Still out of range? Rerun Yes Is issue recurring? Prepare a new dilution of QC and rerun No Issue resolved Yes Issue resolved No Prepare a new dilution of same sample and rerun Still out of range? Yes Is issue recurring? Try new vials of QC, prepare a new dilution and rerun No Issue resolved Yes Issue resolved No Redraw sample or verify by alternate method Still out of range? Yes Verify pipette is calibrated and proper technique followed Yes Is issue recurring? No Issue resolved No Is issue recurring? Yes Go to A Issue resolved Page 1 of 4 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 55 Reticulocyte Troubleshooting for CELL-DYN 3200, 3500 and 3700 RETICULOCYTE TROUBLESHOOTING FOR CELL-DYN 3200, 3500 AND 3700 A Is background on the stain <100? No Rerun background Yes Check maintenance - extended autoclean - weekly - clean metering syringe - clean sheath syringe - drain and fill flow cell Yes Is background <100? No Maintenance up to date? No Try new stain and rerun Redo maintenance and rerun controls No Controls within range? Yes Controls within range? No Do WOC gains and reticulocyte gains match? Yes Yes Issue resolved No Issue resolved Yes 1. 2. 3. Rerun QC Follow procedure below: For WOC gains: go to Main menu, Diagnostics, F12, F1, Set point entry (0, 10, 90), and print screen For reticulocyte gains: go to Reticulocyte mode, Diagnostics, F12, F1, WOC (0, 10, 90) Compare WOC gains and reticulocyte gains (0, 10, 90) Yes Do WOC and reticulocyte gains match? No Go to B Go to C Page 2 of 4 56 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 Reticulocyte Troubleshooting for CELL-DYN 3200, 3500 and 3700 RETICULOCYTE TROUBLESHOOTING FOR CELL-DYN 3200, 3500 AND 3700 B C Copy WOC gains as follows: Enter WOC values (0, 10, 90) into reticulocyte gains screen and press Set analyzer to save entry Rerun patient samples Yes QC within range? Rerun QC No Issue resolved Yes Patient samples OK? Run 3 -- 5 normal patient samples in regular patient mode No Controls within range? No Verify by alternate method Yes Issue resolved Is neutrophil population at 0/10 scattergram at the 2.5 by 2.5 hash marks? Yes Rerun QC No Controls within range? No Dispatch Service Check channel offsets are <2.0 (only for CELL-DYN 3500 and 3700) Yes Issue resolved Go to D Page 3 of 4 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 57 Reticulocyte Troubleshooting for CELL-DYN 3200, 3500 and 3700 RETICULOCYTE TROUBLESHOOTING FOR CELL-DYN 3200, 3500 AND 3700 D Issue resolved Yes Controls within range? Rerun controls Yes Channel offsets within range? (only for CELL-DYN 3500 and 3700) No Drain/fill flow cell, run autoclean, clean WOC mixing chamber (see WOC mixing chamber cleaning procedure) and recheck channel offsets (only for CELL-DYN 3500 and 3700) No Issue resolved Yes Controls within range? Rerun QC No Dispatch Service References: 1. CELL-DYN 3500 Operator's Manual, June 1999: Chapter 14, 14-1 to 14-8 2. CELL-DYN 3700 Operator's Manual, November 2000: Chapter 14, 14-1 to 14-87 3. CELL-DYN 3200 Operator's Manual, October 2001: Chapter 14 Page 4 of 4 58 CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 WOC Flow Cell Cleaning Procedure (CELL-DYN 3000/3500/3700) WOC FLOW CELL CLEANING PROCEDURE (CELL-DYN 3000/3500/3700) Materials/Tools Needed: • 25% Bleach solution • 10 mL syringe Step Action 1 Fill syringe with 25% bleach solution. 2 From the Main Menu, press • • 3 SPECIAL PROTOCOLS DISABLE/ENABLE ANALYZER Insert bleach into WBC/WOC mixing chamber. NOTE: Follow the steps below to avoid damaging the plastic port of mixing chamber on CELL-DYN 3500/3700. a. Follow tubing from mixing chamber to Overflow Chamber. CELL-DYN 3000: top port of WBC mixing chamber CELL-DYN 3500/3700: left top port of WOC mixing chamber b. Disconnect tubing from overflow chamber. c. Connect syringe to tubing. d. Inject bleach solution and fill WBC/WOC mixing chamber 2/3 full. e. Pull back on syringe slightly to remove any excess pressure and liquid trapped in tubing. f. 4 Remove syringe and reconnect tubing to overflow chamber. For Special Protocols Menu, press • ENABLE ANALYZER 5 Perform 2 autocleans with water to allow bleach to enter and clean flow cell. 6 Run 3 background counts. 7 Verify background counts are within specifications. CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101 59 WOC Flow Cell Cleaning Procedure (CELL-DYN 3200) WOC FLOW CELL CLEANING PROCEDURE (CELL-DYN 3200) Materials/Tools Needed: • 25% Bleach solution • 10 mL syringe with tubing attached Step Action 1 Fill syringe with 25% bleach solution. 2 From the Main Menu, press • • 3 SPECIAL PROTOCOLS DISABLE/ENABLE ANALYZER Insert bleach into WBC/WOC mixing chamber. a. Remove tubing from mixing chamber (WBC or RBC/PLT). b. Connect syringe and attach tubing to mixing chamber. c. Inject bleach solution and fill WBC mixing chamber 2/3 full. d. Pull back on syringe slightly to remove any excess pressure and liquid trapped in tubing. e. Remove syringe and reconnect tubing to overflow chamber. 4 For Special Protocols Menu, press • 60 ENABLE ANALYZER 5 Perform 2 autocleans with water to allow bleach to enter and clean flow cell. 6 Run 3 background counts. 7 Verify background counts are within specifications. CELL-DYN® 3000/3500/3700/3200 Troubleshooting Guide 67483-101