EEG 317: Instrumentation and Measurment I Lecture 1: Theory of Errors 1.1 Introduction All branches of engineering perform two main functions: the design of equipment and processes, and the proper operation and maintenance of equipment and processes. These functions require measurement. Hence, the role of measurement cannot be ignored in the engineering field. The measurement of a quantity is fundamentally an act or outcome of comparison between a quantity (whose magnitude is unknown) and a predefined standard. In order that the results of the measurement are meaningful, there are two basic requirements: 1. The standard used for comparison purposes must be accurately defined and should be commonly accepted. 2. The apparatus used and the method adopted must be provable The methods of measurement can be categorized into two, namely, direct methods and indirect methods. The latter is adopted in engineering measurement systems due to the inaccuracies in the direct methods as a result of human factors coupled with their lower sensitivity. Measurements generally involve the use of an instrument as a physical means of determining a quantity or variable. Therefore, we can define an instrument as a device for determining the value or magnitude of a quantity or variable. The historical development of instruments encompasses three phases, namely, mechanical instruments, electrical instruments and electronic instruments. It is important to note that no instrument is perfectly accurate; thus, no measurement is free from error. Assignment 1a - Highlight the main features, advantages, and disadvantages of the aforementioned categories of instruments. 1.2 Types of Errors No measurement, whether conducted in a laboratory or elsewhere, is perfectly accurate. All have some error or inaccuracy. Hence, it is pertinent to determine how different errors have entered into measurement, how they are specified, and how they combine to create even a greater error in measurement systems. Errors may come from different sources and are usually classified under three main headings: 1-1 1-2 Lecture 1: Theory of Errors Gross errors: This class of errors is attributed to human errors, such as misreading of instruments, incorrect adjustment and improper application of instruments, and computational mistakes. Systematic errors: These errors can be further categorized into three, namely instrumental errors, environmental errors, and observational errors. – Instrumental errors: these types of errors occur due to the shortcomings in the instrument, misuse of the instruments, and loading effects of instruments. – Environmental errors: These errors are due to conditions external to the measuring device, including conditions in the area surrounding the instrument. This may be effects of temperature, humidity, dust, vibrations or of external magnetic or electrostatic fields. – Observational errors: There are many sources of observation error. A typical example is the parallax error, which is incurred when reading a measurement from an angle rather than in line with the vision of the observer. Random errors: Sometimes called residual errors are those due to the cause that cannot be directly established because of the random variations in the parameter or the system of measurement. Assignment 1b - Give an elaborate discussion on the aforementioned types of errors and how they can be avoided. In most instruments, the accuracy is guaranteed to be within a certain percentage of fullscale reading. Components are guaranteed to be within a certain percentage of the rated value. Thus, the manufacturer has to specify the deviations from the true or nominal value of a particular quantity. These limits lead us to the theory of absolute and relative treatment of errors. 1.3 Absolute Errors and Relative Errors Given the nominal value of a resistor as 500 Ω with a tolerance of ±5% as depicted in Figure 1.1(a), this means the resistor will have a value between the limits 475 Ω and 525 Ω. Hence, we can say the resistor has a possible error of ±25 Ω. The ±25 Ω is referred to as the absolute error. When the error is expressed as a percentage or as a fraction of the total 1 resistance, it becomes a relative error. Thus the ±25 Ω is ±5%, relative to 500 Ω, or ± 20 of 500 Ω. Note that percentages are usually employed to express errors in resistance and other electrical quantities. The terms accuracy and tolerance are also used. A resistor with a possible error of ±5% is said to be accurate to ±5%, or have a tolerance of ±5%. Mathematically, ρm = ρt ± εa , (1.1) Lecture 1: Theory of Errors 1-3 Figure 1.1: Resistor tolerance is represented by a gold color band. where ρm is the actual or measured value, ρt is the nominal value, and εa is the absolute error. The relative error in fractional and percentage form can be expressed as given in (1.2) and (1.3), respectively. εa (1.2) εr = , ρt Percentage relative error %εr = εr × 100. (1.3) Also, from (1.1), limiting values in terms of relative errors can be expressed as ρm = ρt ± εa = ρt ± εr ρt = ρt (1 ± εr ). (1.4) In addition, when the deviation limits from the nominal value are given, we can take the maximum deviation from the ρm to be the erroneous quantity. Thus we have εa = ρm − ρt . (1.5) Illustration: A component manufacturer constructs a certain resistance to be anywhere between 1.14 kΩ and 1.26 kΩ and classifies them to be 1.2 kΩ resistors. What tolerance should be stated? Solution: From (1.5) εa = 1.26 kΩ − 1.2 kΩ = 0.06 kΩ. (1.6) Thus, εa = ±0.06 kΩ. From (1.2) ±0.06 kΩ = 0.05 1.2 kΩ % εr = 0.05 × 100 = 5%. εr = Take note of the extension of the question in Example 2-1, page 16 of reference [1]. (1.7) (1.8) 1-4 Lecture 1: Theory of Errors 1.4 Accuracy, Precision, Resolution, and Significant Figures 1.4.1 Accuracy and Precision Accuracy refers to the degree of closeness or conformity to the true value of a quantity under measurement. Precision refers to the degree of agreement within a group of measurements or instrument. The term ’Precise’ means sharply defined. Consider the digital voltmeter Figure 1.2: Digital voltmeter with a precision of 1 mV. indication shown in Figure 1.2. For the 8.135 V quantity indicated, the last (right-side) numeral refers to millivolts. If the measured quantity increases or decreases by 1 mV, the reading becomes 8.136 V or 8.134 V, respectively. Therefore, the voltage is measured with precision of 1 mV. Thus we can infer that the measurement precision depends on the smallest change that can be observed in a measured quantity. Suppose the digital voltmeter has an accuracy of ±0.2%. The measured voltage is 8.135 V ±0.2%, or 8.135 ±16 mV, meaning that the actual voltage is somewhere between 8.119 V and 8.151 V. Thus we find that the quantity is measured with a precision of 1 mV, the measurement accuracy is ±16 mV 1.4.2 Resolution The measurement precision of an instrument defines the smallest change in measured quantity that can be observed. This smallest change in measured value to which an instrument will respond is referred to as the resolution of the instrument. In the case of the digital voltmeter depicted in Figure 1.2, the meter can read to a precision of 1 mV, 1 mV is the smallest voltage change that can be observed. Thus, the measurement resolution is 1 mV. Lecture 1: Theory of Errors 1.4.3 1-5 Significant figures The number of significant figures used in a measured quantity indicates the precision of the measurement. For the 8.135 V measurement in Figure 1.2, the four significant figures show that the measurement precision is 0.001 V, or 1 mV. Assuming the precision of 10 mV or 0.01 V, the display would be 8.13 V or 8.14 V; i.e., there would be only three significant figures. The former, with more significant figures, expresses a measurement of greater precision than the latter. Consider the result of using an electronic calculator to determine a resistance value from digital measurement of voltage and current. R= 8.14 V V = = 3.493 562 232 kΩ. I 2.33 mA (1.9) It will be unreasonable to have an answer containing 10 significant figures. The only reasonable approach is to use the same number of significant figures in the answer as in the original quantities. So the calculation becomes R= 1.5 V 8.14 V = = 3.49 kΩ. I 2.33 mA (1.10) Measurement Error Combinations When using two or more instruments to calculate a quantity, it must be assumed that the errors due to the instrument inaccuracy combine in the worst possible way. The resulting error is then larger than the error in any one instrument. This can be easily found by considering the relative increment of the measured value if the final result is in the form of an algebraic equation. 1.5.1 Sum of Quantities Where a quantity is determined as the sum of two measurements, the total error is the sum of absolute errors in each measurement. As illustrated in Figure 1.3(a), E = V1 + V2 = (V1 ± ∆V1 ) + (V2 ± ∆V2 ) = (V1 + V2 ) ± (∆V1 + ∆V2 ) (1.11) Illustration: Three resistors have a nominal resistance of 330 Ω each and are connected in series. One has a ±5% tolerance, and the other two are ±10%. Calculate the maximum and minimum values of the total resistance. 1-6 Lecture 1: Theory of Errors Figure 1.3: Measurement error combinations. (a) Error in sum of quantities equal sum of errors. (b) errors in difference of quantities equals sum of errors. (c) Percentage error in product or quotient of quantities equals sum of percentage errors Solution: Let R1 = 330 Ω ± 5% = 330 Ω ± 16.5 (1.12) R2 = R3 = 330 Ω ± 10% = 330 Ω ± 33 (1.13) RT otal = (330 Ω ± 16.5) + 2(330 Ω ± 33) = 990 ± 82.5 (1.14) max Thus, we have RTmin otal = 907.5 and RT otal = 1072.5 1.5.2 Difference of Quantities Figure 1.3(b) depicts a scenario in which a potential difference is determined as the difference between two measured voltages. Here again, the errors are additive. E = V1 − V2 = (V1 ± ∆V1 ) − (V2 ± ∆V2 ) = (V1 − V2 ) ± (∆V1 + ∆V2 ) (1.15) Lecture 1: Theory of Errors 1.5.3 1-7 Product of Quantities Consider Figure 1.3(c). When a calculated quantity is the product of two quantities, the percentage error is the sum of the percentage errors in each quantity. P = EI = (E ± ∆E)(l ± ∆I) = EI ± E∆I ± I∆E ± ∆E∆I (1.16) P = EI ≃ ±(E∆I + I∆E) (1.17) Since ∆E∆I is very small, E∆I + I∆E × 100% EI ! ∆I ∆E = + × 100% I E Percentage error = % error in P = (% error in I) + (% error in E) 1.5.4 (1.18) (1.19) Quotient of Quantities Following the illustration in Figure 1.3(c), it can be shown that the percentage error is the sum of the percentage errors in each quality. %error in 1.5.5 E = (% error in I) + (% error in E) I (1.20) Quantity Raised to Power % error in AB = B(% error in A) Illustration: An 820 Ω resistor with an accuracy of ±10% carries a current of 10 mA. The current was measured by an analog ammeter on a 25 mA range with an accuracy of ±2% of full scale. Calculate the power dissipated in the resistor, and determine the accuracy of the result. Solution: P = I 2R = (10 mA)2 × 820 Ω = 82 mW Error in R = ±10% (1.21) 1-8 Lecture 1: Theory of Errors The absolute error of the instrument can be computed as εa = εr ρ t = ±2% × 25 = ±0.5 mA (1.22) The magnitude of the current being measured is 10 mA, and the corresponding relative error at this current can be computed as, that is, ±0.5 mA % = ±5% 10 mA (1.23) error in I 2 = 2(±5%) = ±10% (1.24) error in P = (% error in I 2 ) + (% error in R) = ±(10% + 10%) = ±20% (1.25) error in I = 1.6 Basics of Statistic Revise arithmetic mean, deviation, standard deviation, and probable error in Section 2.5 of reference [1] References [1] David A. Bell, Electronic Instrumentation and Measurements, Second Edition. Theory of Errors 1 / 14 Introduction The measurement of a quantity is fundamentally an act or outcome of comparison between a quantity and a predefined standard. The standard used for comparison purposes must be accurately defined and should be commonly accepted. The apparatus used and the method adopted must be provable. Methods of measurement can be categorized into two, namely, direct methods and indirect methods. Measurements generally involve the use of an instrument as a physical means of determining a quantity. An instrument can be defined as a device for determining the value or magnitude of a quantity or variable. Based on historical development, instruments can be categorized under mechanical instruments, electrical instruments and electronic instruments. (Assignment) No instrument is perfectly accurate; thus, no measurement is free from error. 2 / 14 Types of Errors Gross error: This class of errors is attributed to human errors. Systematic errors: These errors can be further categorized into three, namely instrumental errors, environmental errors, and observational errors. Instrumental errors: these types of errors occur due to the shortcomings in the instrument, misuse of the instruments, and loading effects of instruments. ▶ Environmental errors: These errors are due to conditions external to the measuring device, including conditions in the area surrounding the instrument. This may be effects of temperature, humidity, dust etc. ▶ Observational errors: There are many sources of observation error. A typical example is the parallax error. ▶ Random errors: are those due to the cause that cannot be directly established because of the random variations in the parameter or the system of measurement (Assignment). 3 / 14 Absolute Errors and Relative Errors The diagram above illustrates a resistor with a nominal resistance of 500 Ω having a tolerance of ±5% The resistance can also be expressed as 500 ± 25. The ±25 Ω is referred to as the absolute error. When the error is expressed as a percentage or as a fraction of the total resistance, it 1 becomes a relative error i.e., ±5% or ± 20 . 4 / 14 Absolute Errors and Relative Errors Contd. Mathematically, ρm = ρt ± εa , (1) ρm is the actual or measured value, ρt is the nominal value, and εa is the absolute error. The relative error in fractional and percentage form can be expressed as given in (2) and (3), respectively. εr = εa , ρt Percentage relative error %εr = εr × 100. (2) (3) The limiting values in terms of relative errors can be expressed as ρm = ρt ± εa = ρt ± εr ρt = ρt (1 ± εr ). (4) 5 / 14 Absolute Errors and Relative Errors Contd. Illustration: A component manufacturer constructs a certain resistance to be anywhere between 1.14 kΩ and 1.26 kΩ and classifies them to be 1.2 kΩ resistors. What tolerance should be stated? Solution: εa = 1.26 kΩ − 1.2 kΩ = 0.06 kΩ. (5) Thus, εa = ±0.06 kΩ. From (2) ±0.06 kΩ = 0.05 1.2 kΩ % εr = 0.05 × 100 = 5%. εr = (6) (7) 6 / 14 Accuracy, Precision, Resolution, and Significant Figures Accuracy refers to the degree of closeness or conformity to the true value of a quantity under measurement. Precision refers to the degree of agreement within a group of measurements or instrument. The term ’Precise’ means sharply defined. If the measured quantity increases or decreases by 1 mV, the reading becomes 8.136 V or 8.134 V, respectively. Therefore, the voltage is measured with precision of 1 mV. Suppose the digital voltmeter has an accuracy of ±0.2%, what will be the measurement accuracy in terms of absolute error? 7 / 14 This smallest change in measured value to which an instrument will respond is referred to as the resolution of the instrument. The measurement resolution is 1 mV. The number of significant figures used in a measured quantity indicates the precision of the measurement. For the 8.135 V measurement shown in the previous slide, the four significant figures show that the measurement precision is 0.001 V, or 1 mV. Assuming the precision of 10 mV or 0.01 V, the display would be 8.13 V or 8.14 V; i.e., there would be only three significant figures. Which of the measurements expresses a better precision the one with four or three SFs? 8 / 14 Measurement Error Combinations (a) Sum of quantities - (V1 + V2 ) ± (∆V1 + ∆V2 ) Illustration: Three resistors have a nominal resistance of 330 Ω each and are connected in series. One has a ±5% tolerance, and the other two are ±10%. Calculate the maximum and minimum values of the total resistance. 9 / 14 Solution: Let R1 = 330 Ω ± 5% = 330 Ω ± 16.5 R2 = R3 = 330 Ω ± 10% = 330 Ω ± 33 RTotal = (330 Ω ± 16.5) + 2(330 Ω ± 33) = 990 ± 82.5 (8) (9) (10) min = 907.5 and R max = 1072.5 Thus, we have RTotal Total (b) Difference of quantities ⇒ (V1 − V2 ) ± (∆V1 + ∆V2 ) (c) Product of quantities ⇒ Given P = EI , % error in P = (% error in I ) + (% error in E ) Quotient of quantities ⇒ %error in EI = (% error in I ) + (% error in E ) Quantity raised to power ⇒ % error in AB = B(% error in A) 10 / 14 Illustration: An 820 Ω resistor with an accuracy of ±10% carries a current of 10 mA. The current was measured by an analog ammeter on a 25 mA range with an accuracy of ±2% of full scale. Calculate the power dissipated in the resistor, and determine the accuracy of the result. Solution: P = I 2R = (10 mA)2 × 820 Ω = 82 mW (11) Error in R = ±10% The absolute error of the instrument can be computed as εa = εr ρt = ±2% × 25 = ±0.5 mA (12) The magnitude of the current being measured is 10 mA, and the corresponding relative error at this current can be computed as, that is, 11 / 14 ±0.5 mA % = ±5% 10 mA (13) error in I 2 = 2(±5%) = ±10% (14) error in I = error in P = (% error in I 2 ) + (% error in R) = ±(10% + 10%) = ±20% (15) 12 / 14 References I [1] David A. Bell. Electronic Instrumentation and Measurements, Second Edition. Prentice-Hall, 2003. 13 / 14 Electromechanical Instruments 1/9 Permanent Magnet Moving Coil (PMMC) A typical PMMC instrument comprises a lightweight coil of copper wire suspended in the field of a permanent magnet. Current passing through wire causes the coil to generate a magnetic field that interacts with the field from the permanent magnet. The interaction causes a partial rotation of the coil. A pointer connected to the coil deflects over a calibrated scale, indicating the level of current flowing in the wire. Basically, the PMMC is a low-level DC ammeter. However, it can also be made to measure a wider range of direct current and function as a voltmeter or ohmmeter. Figure 1: DC/AC moving coil ammeter 2/9 Fundamentals of Deflection Instruments In PMMC instruments, three operating forces are involved for a pointer deflection to occur over a calibrated scale, namely deflection force, controlling force, and damping force. Deflection force The deflection force is required for moving the pointer from its zero position when a current flows. It occurs due to the interaction between the magnetic field produced by the coil pivoted between the poles of the permanent magnet. As a result, a force is exerted on the coil turns, causing the coil to rotate on its pivots . Figure 2: The deflecting force in a PMMC instrument. 3/9 Fundamentals of Deflection Instruments Contd. Controlling force In the absence of a controlling force, the pointer will swing beyond the final position for any magnitude of current, causing the deflection to be indefinite. The controlling force provided by spiral springs produces a force equal and opposite to the deflecting force at the final steady position of the pointer to make the pointer deflection definite for a particular magnitude of current. Also, the controlling force brings back the coil and pointer to their zero position when the deflecting force is removed. Figure 3: Controlling force from spiral spring 4/9 Fundamentals of Deflection Instruments Contd. Damping force The deflection and controlling forces are produced by systems that have inertia; hence, the pointer and the coil tend to oscillate for some time before settling down at their final position. - Fig. 4(a) A damping force is required to damp out the oscillations. This type of force must be present only when the coil is in motion; thus, it must be generated by the motion of the coil. In PMMC instruments, the damping force is normally provided by eddy currents that is induced from the coil frame made of aluminum. - Fig. 4(b) The eddy currents induced in the coil frame set up a magnetic flux that opposes the coil motion, thus damping the oscillations of the coil. Figure 4 5/9 Assignment Figure 5: PMMC construction. Explain in detail the construction of a PMMC instrument shown in Figure 5. 6/9 Torque Equation and Scale. As illustrated in Fig. 6, a force F is exerted on each side of the coil when a current flows through the coil. This force can be described by (1) F = BIL newtons, (1) B is the magnetic flux density in tesla, I is the current in amperes, and L is the length of the coil in meters. Given N turns of coil; the force acts on the two sides of the coil, thus, the total force can be given as F = 2BILN newtons. (2) Figure 6 The force on each side acts at radius r, producing a deflecting torque. TD = 2BILN r, Newton meters = BILN (2r), TD = BLIN D, (3) (4) (5) where D is the coil diameter. 7/9 The controlling torque TC is proportional to the actual angle of the deflection of the pointer. TC = Kθ, (6) Note, the expression in (8) indicates that the pointer deflection is always proportional to the coil current and the scale is linear. Illustration where K is a constant. For a A PMMC instrument has a 0.12 T magnetic flux density given deflection, the controlling in its air gaps. The coil dimensions are D = 1.5 cm and and deflecting torques are equal: L = 2.25 cm. Determine the number of coil turns required to give a torque of 4.5 µN.m when the coil Kθ = BLIN D. (7) current is 100 µA. From (7), all quantities are constant except I and θ, θ = CI, N= (8) 4.5 × 10−6 TD = BLID 0.12 × 1.5 × 10−2 × 100 × 10−6 × 2.25 ≈ 1111 where C is a constant. 8/9 Galvanometer A galvanometer is essentially a PMMC instrument of a center-zero scale type, designed to be sensitive to extremely low current levels (microamperes). The pointer can deflect either right or left of zero, depending on the direction of the current through the moving coil. Galvanometers are often employed to detect zero current or voltage in a circuit rather than to measure the actual level of current or voltage. Galvanometers find their principal application in bridge and potentiometer measurements, where their function is to indicate zero current. A galvanometer must be protected from excessive current flow that might occur when the voltage across the instrument terminal is not close to zero. Figure 7 Protect is provided by an adjustable resistance connected in shunt with the instrument. 9/9 PMMC Instruments Contd 1/8 DC Ammeter Ammeter circuit Ammeters are connected in series with a circuit in which current is to be measured. They should have resistance much lower than the circuit resistance in so that they cause a small voltage drop and consequently absorb small power. However, maximum pointer deflection is produced by a very small current, and the coil is usually wound of thin wire that would be quickly destroyed by large currents. When heavy currents are to be measured, the major part of the current is bypassed through a low resistance called shunt. Figure 1: Ammeter circuit The shunt Rs is connected in parallel with the instrument coil as depicted in Figure 1. Im , Is and Rm are the meter current, shunt current, and coil/meter resistance, respectively. 2/8 Illustration 0.05 mA × 99 Ω Vm An ammeter depicted in Figure 1, has a = = 4.95 mA Is = R 1 Ω s PMMC instrument with a coil resistance of Rm = 99 Ω and FSD current of 0.1 mA, shunt Thus, I = (0.05 + 4.95) mA = 5 mA resistance Rs = 1 Ω. Determine the total What is the value of I at 0.25 FSD? current passing through the ammeter at (a) ✓ Going by the above illustration, the meter FSD and (b) 0.5 FSD and (b) 0.25 FSD. can be calibrated to indicate 10 mA. Solution I = Im + Is 0.1 mA × 99 Ω Vm = = 9.9 mA Is = Rs 1Ω Thus, I = (0.1 + 9.9) mA = 10 mA At 0.5 FSD Illustration - Shunt Resistance A PMMC instrument with a 750 Ω coil resistance gives FSD with a 500 µA coil current. Determine the required shunt resistance to convert the instrument into a dc ammeter with an FSD of (a) 50 mA and (b) 30 mA Im = 0.5 × 0.1 mA = 0.05 mA 3/8 Solution Vm = Im Rm = 750 Ω × 500 µA = 375 mV Is = I − Im = 50 mA − 500 µA = 49.5 mA Vm 375 mV Rs = = Is 49.5 mA = 7.576 Ω Swamping Resistance: Errors can be introduced in the ammeter measurements due to resistance change in the coils as a result of change in temperature of the coil. Figure 2: Ammeter circuit with swamp resistance To alleviate this effect, a swamping resistance made of manganin or constantan is connected in series with the coil as shown in Fig . If the swamping resistance is nine times the coil resistance, a 1% change in coil resistance would result in a total resistance change of 0.1%. 4/8 Multirange Ammeter: The circuit development of a multirange ammeter can be realized as depicted in Figures 3 and 4. A critical safety factor considered in the two designs is that the instruments are not left without a shunt in parallel with it even for a brief instance. Figure 4: Multirange ammeter - Config. 2 Figure 3: Multirange ammeter - Config. 1 In Fig. 4, the switch is at contact B, and the resistors R1 + R2 + R2 which constitute the ayrton shunt, are in parallel with the instrument. The meter circuit remains Rm 5/8 Solution Switch at contact B When the switch is at contact C, the resistor R3 is in series with the meter, and Vs = Im Rm = 100 µA × 1.2 kΩ R1 + R2 is in parallel with Rm + R3 = 0.12 V Similarly, with the switch at contact D, Vs 0.12 V Is = = R1 is in parallel with Rm + R2 + R3 . (R1 + R2 + R3 + R4 ) 0.4 Ω Hence, the instrument is never left = 0.3 A without a parallel-connected shunt. I = Im + Is = 100 µA + 0.3 A = 300.1 mA Illustration - Multirange ammeter A dc ammeter consists of an Aryton shunt in parallel with a PMMC instrument that has a 1.2 kΩ coil resistance and 100 µA FSD. The Aryton shunt is made up of four 0.1 Ω series-connected resistors. Calculate the ammeter range at each setting of the shunt. ≃ 300 mA Switch at contact E Vs = Im (Rm + R2 + R3 + R4 ) = 120.03 mV Vs 120.03 mV Is = = = 1200.3 mA R1 0.1 Ω I = 100 µA + 1200.3 A = 1200.4 mA ≃ 1200 mA 6/8 DC Voltmeter Voltmeter circuit: The deflection of a PMMC instrument is proportional to the current flowing through the moving coil. The coil current is directly proportional to the voltage across it. Hence, the scale of PMMC instrument could be calibrated to indicate voltage. The voltmeter range can be extended by connecting a resistance in series with the instrument as shown in Fig. 5. Because it increases the range of the voltmeter, the multiplier resistance, the series resistance is termed a multiplier resistance. Figure 5: Multirange voltmeter Illustration A PMMC instrument with 900 Ω coil resistance and FSD of 75 µA is to be used as a dc voltmeter. Calculate the individual multiplier resistance to give an FSD of (a) 100 V (b) 30 V, and (c) 5 V. 7/8 Solution At 100 V V = Im (Rm + R1 ) 100 V V − Rm = − 900 Ω R1 = Im 75 µA = 1.33 MΩ Rectifier Voltmeter 8/8 PMMC Instruments Contd 1 / 10 Quiz In the circuit illustrated above, all the diodes are of silicon, each with a forward resistance of 2 Ω. Calculate the current through the 50 Ω resistor. 2 / 10 Rectifier Voltmeter Full-Wave (FW) Rectifier Voltmeter: The FW rectifier voltmeter as depicted in Figure 1, uses a series multiplier resistor to limit the current flow though the PMMC instrument. The meter deflection is proportional to the average current, which is 0.637 times peak current. → Idc = Iav = 2Iπm However, the actual current (or voltage) to be indicated in ac measurement is normally the rms quantity, which is 0.707 of the peak value or 1.11 times the Im average value. → Irms = √ 2 Since there are direct relationships between rms, peak, and average values, Figure 1: Full-Wave rectifier voltmeter the meter scale can be calibrated to indicate rms volts. Illustration: A PMMC instrument with F SD = 100 µ and Rm = 1 kΩ is to be employed as an ac voltmeter with F SD = 100 V (rms). Silicon diodes are used in the bridge rectifier circuit of Figure 1. 3 / 10 Calculate the multiplier resistance value required. Solution: Applying KCL in Fig. 2, we have − Vm + Im Rs + VD1 + Im Rm + VD4 = 0 Im (Rs + Rm ) = Vm − (VD1 + VD4 ) Vm − (VD1 + VD4 ) Rs = − Rm Im Vm = Vrms × √ Figure 2 2 = 100 V × 1.414 = 141.4 V VD1 = VD4 = 0.7 Iav 100 µA Im = = = 157 µA 0.637 0.637 141.4 V − 1.4 V Rs = − 1 kΩ = 890.7 kΩ 157 µA ✓ Rectifier Ammeter: ✓ Series Ohmmeter: Assignment: Problems 3.23 and 3.24, page 94 of [1]. Submission date: 22 January 2025, 9:00 AM. 4 / 10 References I [1] David A. Bell. Electronic Instrumentation and Measurements, Second Edition. Prentice-Hall, 2003. 5 / 10 Analog Electronic Instruments 6 / 10 Introduction Voltmeters constructed of moving-coil instrument and multiplier resistor have some limitations. They cannot measure very low voltages due to loading effect, and their resistance is too low for measurements in high-impedance circuitry. The analog electronic voltmeter configuration include the ✓ Emitter-Follower Voltmeter ✓ FET-Input Voltmeter These limitations are overcome by the use of electronic circuit that offers high input resistance, and which amplify low voltages to measurable levels. Electronic voltmeter can be analog instruments, in which the measurement is indicated by a pointer over a calibrated scale, or digital instruments which displays the measurement in numerical form. Figure 3: Emitter-Follower Voltmeter 7 / 10 Introduction The basic emitter-follower voltmeter circuit depicted in Fig. 3, shows a PMMC instrument and a multiplier resistance Rs connected in series with the transistor emitter. The DC supply is connected - positive to the transistor collector and negative to the deflection meter. The positive terminal of voltage E to be measured is supplied to the transistor base, and its negative is connected to the same terminal as the power supply negative. The transistor base current is substantially lower than the meter current. IB ≈ Im , hF E where hF E is the transistor current gain. Thus the circuit input resistance is Ri ≈ E , IB which is much larger than the meter circuit resistance (Rs + Rm ) Illustration A simple emitter-follower circuit voltmeter shown in Fig. 3 has VCC = 12 V, Rm = 1 kΩ, a 2 mA meter, and a transistor of with hF E = 80. Calculate a suitable resistance for Rs to give full-scale deflection when E = 5 V. Also, determine the voltmeter input resistance. 8 / 10 Solution: Applying KCL in to the left loop of Fig. 2, we have −E + VBE + Im (Rm + Rs ) = 0 E − VBE − Im (Rm ) Rs = Im 5 V − 0.7 V − (2 mA)(1 kΩ) = = 1.15 kΩ 2 mA Im 2 mA = = 25 µA hF E 80 E 5V = 200 kΩ Ri = = IB 25 µA IB = The voltage drop VBE introduces an error in the simple emitter-follower voltmeter. Figure 4: Modified Emitter-Follower Voltmeter The error can be eliminated using a potential divider and an additional emitter follower as depicted in Fig. 4. When E = 0 V, the base voltage of Q2 is adjusted to give zero meter current. 9 / 10 This makes VP = 0 V, VE1 = VE2 = −0.7 V, and V = 0 V. Hence the voltage drop is removed Assignment: Sketch the complete circuit of an emitter follower voltmeter using a FET stage. Carefully explain the circuit operation Submission date: To be determined later. 10 / 10 Quiz In the circuit above, R1 = 600 Ω, R2 = 1000 Ω, R3 = 400 Ω, R4 = 1000 Ω, and R5 = 500 Ω. Determine the measurement error caused by the loading effect in the voltmeter when the voltage across AB is measured. Take the voltmeter internal resistance as 9500 Ω. 1/8 Digital Instruments 2/8 Introduction While the analog instrument display the quantity to be measured in terms of deflection of a pointer, digital instruments the measurement value in form of a decimal number. The digital meter works based on the principle of quantization. For illustration purpose, the operating principles of digital voltmeter (DVM) (Ramp-type and dual-slope-integrator DVMs) will be considered. A DVM essentially consists of analog-to-digital converter (ADC) with a set of seven-segment numerical displays to indicate the measured voltage. See Fig 1. Ramp-Type DVMs The ramp-type VDM measures unknown voltage using a ramp signal, based on the time it takes for the ramp to rise from a reference voltage to the input voltage. The input voltage Vi and the ramp signal Vr are compared in the comparator. When Vi ≥ Vr , V1 = 1, the gate is in an open state and the decade counter counts the pulses generated from the clock circuit. When Vi < Vr , V1 = 0, the gate is closed. The output of the decade counter corresponds to the number of clock pulses, which is proportional to the duration of the ramp transition. 3/8 Figure 1: A ramp-type DVM. (a) DVM system. (b) DVM waveforms 4/8 Dual-slope-integrator VDM The latch locks the counter’s output at the end of measurement, storing the count representing the voltage for a stable and accurate display. A BCD to seven-segment driver converts a BCD input into signals that control a seven-segment display. Its function is to translate the 4-bit BCD code (representing decimal digits 0–9) into appropriate outputs Limitation of ramp-type VDM The ramp-type VDMs require precise ramp voltages and precise time periods, both of which can be difficult to maintain. The dual-slope-integrator (DSI) VDM, illustrated in Fig 2 eliminates the limitations of ramp-type VDM by using a special type of ramp generator circuit (or integrator) The integrator capacitor is first charged from the analog input voltage and then discharged at a constant rate to give a time period that is measured digitally. The control waveform for the integrator is derived from the clock generator by use of a frequency divider During time t1 the integrator capacitor is charged to negatively from Vi , giving a negative-going ramp. 5/8 Figure 2: A dual-slope-integrator DVM. (a) System block diagram. (b) System waveforms 6/8 The voltage V0 produced during this phase is directly proportional to Vi The constant current source is then switched into the circuit to discharge the capacitor, thus producing a positive going ramp voltage. The zero voltage comparator is a voltage comparator that gives a high output when the integrator output waveform is negative, and low output at the end of the positive-going ramp. The AND gate has high inputs from both the zero-crossing detector and the control waveform only during the positive-going ramp time, i.e., during t2 Pulses from the clock generator pass through the AND gate to the counting circuits during this time. The counting circuit is reset to zero by the positive-going edge of the control wave at the commencement of t2 , so the output of the counting circuit is a digital measurement of time t2 Since t2 is directly proportional to V0 , and V0 is directly proportional to Vi , the output is a digital measurement of the analog input. 7/8 References I [1] David A. Bell. Electronic Instrumentation and Measurements, Second Edition. Prentice-Hall, 2003. 8/8 Digital Frequency Meter Accurate frequency measurement is essential for precise control and monitoring of systems and devices, ensuring they operate correctly by maintaining the proper oscillation rate or cycles per second. The number of counted cycles per unit time indicates the signal frequency. Assignment: Explain in detail the operation of a digital frequency meter depicted in Figure 1. This is particularly crucial in applications Submission Due Date: such as communication systems, power grids, and electronic circuits, where stable 14 February 2025 at 11:55 PM frequencies are necessary for signal DVM Illustration: Calculate the maximum time integrity and efficient performance. t1 for the digital voltmeter in Figure 1 of To measure the frequency of a certain Lecture 5 presentation. If the clock generator periodic signal, the waveform of that frequency is 1.5 MHz. Also suggest a suitable signal is used to toggle a counter for a frequency for the ramp generator. certain fixed time. 1/2 Figure 1: Digital frequency meter. (a) Frequency meter system (b) System waveforms 2/2