Technical EuropeanService Sales Meeting Training Barcelona 2002 The Amazing ADAPTIV For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 • Biphasic defibrillation • The differences • Update the studies • The damage • The AHA guidelines • Predictions For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Biphasic Defibrillation • Current flows in two phases, first in one direction from one electrode, and then current flows the other way from the other electrode For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Biphasic Defibrillation • How does it work? We really don’t know (it’s magic!) • Is it really better? – It has been in the hospital • EP Lab • AF patients – OOH • No more survivors with biphasic (yet) • Better at certain endpoints (5 sec. post shock, ROSC) For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Defibrillation Electrical Current Stops VF •Must overcome resistance (impedance) to flow Shock VF Defibrillation NSR For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Defibrillation Impedance 60 Current variance due to impedance, energy held constant Peak Current (amps) 50 40 30 20 10 0 25 50 75 100 125 150 Patient Impedance (ohms) The current a heart receives from a 200J shock depends on the patient’s impedance For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Defibrillation Electricity and the Heart • Defibrillators use energy to deliver the shock – Energy is expressed in joules Energy = Current x Voltage x Time Joules = Amps x Volts x Seconds For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 The Differences Histogram of patient impedances Percentage of Patients 16 14 12 10 8 6 4 2 0 30 40 50 60 70 80 90 100 110 120 130 140 Impedance (ohms) Source: Medtronic Physio-Control: Impedance data on 723 SCA patients. For training purposes only 150 More Technical EuropeanService Sales Meeting Training Barcelona 2002 Biphasic Waveforms 40 Monophasic Peak Current Current (amps) 30 Biphasic Peak Current 20 10 0 -10 -20 0 5 10 15 20 Time (msec) Biphasic waveforms have a lower peak current For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Biphasic Current Duration Duration adjusted biphasic waveforms Low Impedance Shock Medium High 50 ohms 75 ohms Current (amps) 125 ohms Time (msec) Biphasic waveforms duration can adjust for impedance For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Voltage Compensation 2000 Voltage Waveforms for 200 J Shocks 1500 50 ohms 75 ohms 100 ohms 125 ohms 150 ohms Voltage (V) 1000 500 0 0 5 10 15 20 -500 -1000 Time (ms) -1500 • Voltage (pressure) can adjust for impedance For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Medtronic ADAPTIV BTE Waveform 40 50 ohms (369J) 75 ohms (356J) 125 ohms (320J) 35 30 Current (amps) 25 20 15 10 5 0 -5 -10 -15 -20 -2 0 2 4 6 8 10 12 14 16 Time (msec) Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors. For training purposes only 18 20 Technical EuropeanService Sales Meeting Training Barcelona 2002 Heartstream Basic Claims • Gives just the right amount to the patient • Low energy--doesn’t damage the heart • Proven technology—most published studies • Only one with AHA rating For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 First shock setting (AED mode), 75 ohms 35 MDS - 208 J delivered 30 Heartstream - 152 J delivered Current (A) 25 Physio - 203 J delivered 20 Zoll - 143 J delivered 15 10 5 0 -5 -10 -15 -2 0 2 4 6 8 10 12 14 Time (ms) For training purposes only 16 18 20 Technical EuropeanService Sales Meeting Training Barcelona 2002 CL6815-00 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Heartstream--The Studies • There are lots of them! What were the results? – Worst success in EP lab (86% versus MPC 100% at 200J biphasic) – Worst results of all manufacturers for AF (Page, et al, CL 6864-00) – 78% at 150J – 90% at 200J versus 97% for MPC • OOH • Never have shown an increase in survival….. but showed better success at other endpoints • Defined shock success at 5 seconds post shock For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Out-of-Hospital Cardiac Arrest Studies Defining outcomes for pre-hospital defibrillation Failed shock: VF present 5 seconds after the shock VF Successful shock: asystole or organized rhythm for 5 seconds after the shock organized For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Out-of-Hospital Cardiac Arrest Studies Defining outcomes for pre-hospital defibrillation – the controversy Success? For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Heartstream--The Studies • OOH study (ORCA or Schneider study) was instrumental in getting Class IIa rating in international guidelines – Compared 150J BTE to escalating MTE • MTE believed to be an inferior waveform For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 A Comparison of Clinically Used Biphasic Waveforms for External Defibrillation • Publication: RG Walker, Academic Emergency Medicine 2001;8(5):432-3 (CL6873-00) For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Methods • Short duration (30 sec), electrically-induced VF • Used each manufacturer’s recommended /default 3-shock protocol for automated defibrillation • 8 pigs; 16 trials for each waveform For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Compared Efficacy of Four Biphasic Defibs: Low Impedance Episodes 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 3 Shocks 2 Shocks 1 Shock HS PC SL Comparison in 8 pigs, 256 VF episodes Impedance 40 +/- 7 ohms Walker RG, et al. Acad Emerg Med. 2001;8(5):432-433. For training purposes only ZL Technical EuropeanService Sales Meeting Training Barcelona 2002 Typical distribution of patient impedances for OOH Ttransthoracic Defibrillation Percentage of Patients Pigs = ~40 ohms Average human = ~ 80 - 85 ohms 16% 14% 12% 10% 8% 6% 4% 2% 0% 30 40 50 60 70 80 90 100 110 120 130 140 150 More Impedance (ohms) For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Compared Efficacy of Four Biphasic Defibs: Medium Impedance Episodes 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 2 Shocks 1 Shock HS PC 3 Shocks SL ZL Comparison in 8 pigs, 256 VF episodes Impedance 92 +/- 7 ohms (50 ohms added) Walker RG, et al. Acad Emerg Med. 2001;8(5):432-433. For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 ZOLL—Main Claims • Has low peak current which causes less damage • Can achieve what MPC does but with less energy (ZOLL’s energy fluctuates a lot) • Lots of published studies support low energy – 2 published studies, one EP and one AF study • Very weak OOH abstract just presented at a conference • Often give Heartstream studies to customers • New, potentially damaging study just presented in May, 2002 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 First shock setting (AED mode), 75 ohms 35 MDS - 208 J delivered 30 Heartstream - 152 J delivered Current (A) 25 Physio - 203 J delivered 20 Zoll - 143 J delivered 15 10 5 0 -5 -10 -15 -2 0 2 4 6 8 10 12 14 Time (ms) For training purposes only 16 18 20 Technical EuropeanService Sales Meeting Training Barcelona 2002 High Impedance 18 LIFEPAK 12 biphasic device (360J) FirstSave (high) M Series (200J) ForeRunner (150J) 16 14 12 Current (amps) 10 8 6 4 2 0 -2 -4 -6 -8 -10 -12 -2 0 2 4 6 8 10 12 14 16 Time (msec) Maximum shock into 125 ohms Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors. For training purposes only 18 20 22 Technical EuropeanService Sales Meeting Training Barcelona 2002 Current Duration 25 20 Shock comparison at maximum energy level High impedance condition (125 ohms) Medtronic Physio-Control (360J) Current (amps) 15 Company B (200J) 10 Increased Duration for Increased Impedance 5 0 0 2 4 6 8 10 12 14 16 18 20 All waveforms generated using actual devices -5 -10 -15 Time (msec) High impedance patients receive more energy with the Medtronic Physio-Control waveform For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 ZOLL Waveform Delivers More Energy and Higher Peak Current at 100J than MPC 100 J setting, 100 ohms 100 J setting, 75 ohms 15 5 0 -5 5 0 -5 -10 -10 -15 -15 -2 0 2 4 6 8 10 12 14 16 18 20 Physio - 102 J delivered Zoll - 119 J delivered 10 Current (A) 10 Current (A) 15 Physio - 102 J delivered Zoll - 120 J delivered -2 0 2 4 6 8 10 Time (ms) Time (ms) For training purposes only 12 14 16 18 20 Technical EuropeanService Sales Meeting Training Barcelona 2002 ZOLL Waveform Delivers More Energy and More Current at 200J than MPC 200 J setting, 75 ohms 25 25 Physio - 203 J delivered Zoll - 224 J delivered 20 15 15 10 10 5 0 -5 5 0 -5 -10 -10 -15 -15 -20 -20 -2 0 2 4 6 8 10 12 14 16 18 20 Physio - 205 J delivered Zoll - 220 J delivered 20 Current (A) Current (A) 200 J setting, 100 ohms -2 0 2 4 Time (ms) 6 8 10 Time (ms) For training purposes only 12 14 16 18 20 Technical EuropeanService Sales Meeting Training Barcelona 2002 Damage • Clinically significant damage has not been proven with monophasic shocks at traditional energy levels • Our ADAPTIV biphasic shocks give 40% less peak current than monophasic shocks For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Damage: None at Levels up to 360J Monophasic Cardiac troponin I levels after monophasic shocks13 Not Elevated Minimally elevated (0.8–1.5 ng/ml)* * Permanent cardiac injury occurs at levels greater than 7.7ng/ml For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 1st Shock Comparison, Average Impedance 25 LIFEPAK 12 biphasic device (200J) FirstSave (low) M Series (120J) ForeRunner (150J) 20 Current (amps) 15 10 5 0 -5 -10 -15 -2 0 2 4 6 8 10 Time (msec) First shock into 75 ohms Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors. For training purposes only 12 14 16 Technical EuropeanService Sales Meeting Training Barcelona 2002 The Illusion of Damage 18 LIFEPAK 12 biphasic device (360J) FirstSave (high) M Series (200J) ForeRunner (150J) 16 14 12 Current (amps) 10 8 6 4 2 0 -2 -4 -6 -8 -10 -12 -2 0 2 4 6 8 10 12 14 16 Time (msec) Maximum shock into 125 ohms Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors. For training purposes only 18 20 22 Technical EuropeanService Sales Meeting Training Barcelona 2002 Summary: Damage • Peak current, not energy causes damage • No documented clinically significant damage with monophasic waveforms • ADAPTIV biphasic waveform at 360J offers peak current levels similar to monophasic at 100J For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Summary: Heartstream • Heartstream – Extends waveform duration – Energy remains fixed at 150 J • Gives most current to low impedance patients • Gives least current to high impedance patients For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Summary: ZOLL • ZOLL – Does good job of delivering higher level of current to high impedance patients – Does not ever extend duration – Must go to BTE waveform at 200J and above 85 ohms • But does not extend duration – Tries to pass off Heartstream’s studies as their own – U. Mass AF study will be discussed Thursday at lunch time session For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 The Guidelines • ILCOR – What is their position? For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 International Guidelines 2000 Regarding 1 Biphasic Defibrillation “Current research confirms that biphasic shock energies of 200J are safe and effective.” (Class IIa) P I-90 “Research has not yet determined the optimal biphasic waveform.” P I-90 American Heart Association/International Liaison Committee on Resuscitation, pp I-90–I-91. For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 The Guidelines on Biphasic Energy Levels “We cannot make a definitive recommendation for the energy for first and subsequent nonescalating biphasic defibrillation attempts.” P I-91 • “Even though both escalating or nonescalating biphasic defibrillators are available, there was insufficient data to recommend one over the other.” Source: Guidelines 2000 for Cardiovascular Resuscitation and Emergency Cardiovascular Care.Circulation. 2000;102(suppl I) 8. August 22, 2000 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Biphasic 350 300 250 200 Monophasic 150 Biphasic 100 50 0 Spontaneous VF Electrically Induced Spontaneous VF Electrically Induced VF For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Author’s Conclusions “These results suggest that clinical studies that involve defibrillating spontaneous arrhythmias in either a prehospital or intensive care unit setting are necessary to determine optimal defibrillation energy levels for defibrillation using BTE waveforms and should include shock strengths to 360 J.” For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Biphasic Defibrillation-MDT 1st OOH Results—46 VF patients • MDS in LIFEPAK 12 – 28 VF patients – Collapse to shock = 13.03 minutes – First shock success 19/28 (68%) – ROSC 15/28 (54%) – Survival 5/28 (18%) • BTE in LIFEPAK 12 – 18 VF patients – Collapse to shock = 12.25 minutes – First shock success 15/18 (83%) – ROSC 15/18 (83%) – Survival 8/18 (44%) Doets, vanAlem, Koster presentation at the World Congress in Australia, May 2002 For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Reality • LIFEPAKs have – Current Duration – Voltage Compensation PLUS – Optional energy settings • Lower risk of refibrillation • Higher second shock success • Wider range of therapy • Trend toward more survivors – Goal: to save more lives For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Predictions • More studies will show biphasic is better in OOH setting • Customers will become more interested (and confused) in biphasic defibrillation For training purposes only Technical EuropeanService Sales Meeting Training Barcelona 2002 Predictions • More manufacturers will try to build devices capable of >200J • The Guidelines will remain open to interpretation • More lives will be saved For training purposes only