Uploaded by marcocg.05

Defibrillation Technology: Biphasic Waveforms & ADAPTIV

advertisement
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
The Amazing
ADAPTIV
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
• Biphasic defibrillation
• The differences
• Update the studies
• The damage
• The AHA guidelines
• Predictions
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Biphasic Defibrillation
• Current flows in two phases, first in one
direction from one electrode, and then current
flows the other way from the other electrode
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Biphasic Defibrillation
• How does it work?
We really don’t know (it’s magic!)
• Is it really better?
– It has been in the hospital
• EP Lab
• AF patients
– OOH
• No more survivors with biphasic (yet)
• Better at certain endpoints (5 sec. post shock, ROSC)
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Defibrillation
Electrical Current Stops VF
•Must overcome resistance (impedance) to flow
Shock
VF
Defibrillation
NSR
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Defibrillation
Impedance
60
Current variance due
to impedance, energy
held constant
Peak Current (amps)
50
40
30
20
10
0
25
50
75
100
125
150
Patient Impedance (ohms)
The current a heart receives from a 200J
shock depends on the patient’s impedance
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Defibrillation
Electricity and the Heart
• Defibrillators use energy to deliver
the shock
– Energy is expressed in joules
Energy = Current x Voltage x Time
Joules =
Amps x Volts x Seconds
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
The Differences
Histogram of patient impedances
Percentage of Patients
16
14
12
10
8
6
4
2
0
30
40
50
60
70
80
90
100
110
120
130
140
Impedance (ohms)
Source: Medtronic Physio-Control: Impedance data on 723 SCA patients.
For training purposes only
150
More
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Biphasic Waveforms
40
Monophasic Peak Current
Current (amps)
30
Biphasic Peak Current
20
10
0
-10
-20
0
5
10
15
20
Time (msec)
Biphasic waveforms have a lower peak current
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Biphasic Current Duration
Duration adjusted biphasic waveforms
Low Impedance Shock
Medium High
50 ohms
75 ohms
Current (amps)
125 ohms
Time (msec)
Biphasic waveforms duration can
adjust for impedance
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Voltage Compensation
2000
Voltage Waveforms
for 200 J Shocks
1500
50 ohms
75 ohms
100 ohms
125 ohms
150 ohms
Voltage (V)
1000
500
0
0
5
10
15
20
-500
-1000
Time (ms)
-1500
• Voltage (pressure) can adjust for impedance
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Medtronic ADAPTIV BTE Waveform
40
50 ohms (369J)
75 ohms (356J)
125 ohms (320J)
35
30
Current (amps)
25
20
15
10
5
0
-5
-10
-15
-20
-2
0
2
4
6
8
10
12
14
16
Time (msec)
Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors.
For training purposes only
18
20
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Heartstream Basic Claims
• Gives just the right amount to the patient
• Low energy--doesn’t damage the heart
• Proven technology—most published studies
• Only one with AHA rating
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
First shock setting (AED mode), 75 ohms
35
MDS - 208 J delivered
30
Heartstream - 152 J delivered
Current (A)
25
Physio - 203 J delivered
20
Zoll - 143 J delivered
15
10
5
0
-5
-10
-15
-2
0
2
4
6
8
10
12
14
Time (ms)
For training purposes only
16
18
20
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
CL6815-00
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Heartstream--The Studies
• There are lots of them! What were the results?
– Worst success in EP lab (86% versus MPC 100% at 200J
biphasic)
– Worst results of all manufacturers for AF (Page, et al, CL
6864-00)
– 78% at 150J
– 90% at 200J versus 97% for MPC
• OOH
• Never have shown an increase in survival….. but showed better
success at other endpoints
• Defined shock success at 5 seconds post shock
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Out-of-Hospital Cardiac Arrest Studies
Defining outcomes for pre-hospital defibrillation
Failed shock: VF present 5 seconds after the shock
VF
Successful shock: asystole or organized rhythm for 5 seconds after the
shock
organized
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Out-of-Hospital Cardiac Arrest Studies
Defining outcomes for pre-hospital defibrillation – the controversy
Success?
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Heartstream--The Studies
• OOH study (ORCA or Schneider study) was
instrumental in getting Class IIa rating in
international guidelines
– Compared 150J BTE to escalating MTE
• MTE believed to be an inferior waveform
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
A Comparison of Clinically Used Biphasic
Waveforms for External Defibrillation
• Publication:
RG Walker, Academic Emergency Medicine
2001;8(5):432-3 (CL6873-00)
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Methods
• Short duration (30 sec), electrically-induced
VF
• Used each manufacturer’s recommended
/default 3-shock protocol for automated
defibrillation
• 8 pigs; 16 trials for each waveform
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Compared Efficacy of Four Biphasic Defibs:
Low Impedance Episodes
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
 3 Shocks
 2 Shocks
1 Shock
HS
PC
SL
Comparison in 8 pigs, 256 VF episodes
Impedance 40 +/- 7 ohms
Walker RG, et al. Acad Emerg Med. 2001;8(5):432-433.
For training purposes only
ZL
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Typical distribution of patient impedances for OOH
Ttransthoracic Defibrillation
Percentage of
Patients
Pigs = ~40 ohms
Average human = ~ 80 - 85 ohms
16%
14%
12%
10%
8%
6%
4%
2%
0%
30 40 50 60 70 80 90 100 110 120 130 140 150 More
Impedance (ohms)
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Compared Efficacy of Four Biphasic Defibs:
Medium Impedance Episodes
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
 2 Shocks
1 Shock
HS
PC
 3 Shocks
SL
ZL
Comparison in 8 pigs, 256 VF episodes Impedance 92 +/- 7 ohms (50 ohms added)
Walker RG, et al. Acad Emerg Med. 2001;8(5):432-433.
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
ZOLL—Main Claims
• Has low peak current which causes less damage
• Can achieve what MPC does but with less energy
(ZOLL’s energy fluctuates a lot)
• Lots of published studies support low energy
– 2 published studies, one EP and one AF study
• Very weak OOH abstract just presented at a conference
• Often give Heartstream studies to customers
• New, potentially damaging study just presented in May, 2002
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
First shock setting (AED mode), 75 ohms
35
MDS - 208 J delivered
30
Heartstream - 152 J delivered
Current (A)
25
Physio - 203 J delivered
20
Zoll - 143 J delivered
15
10
5
0
-5
-10
-15
-2
0
2
4
6
8
10
12
14
Time (ms)
For training purposes only
16
18
20
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
High Impedance
18
LIFEPAK 12 biphasic device (360J)
FirstSave (high)
M Series (200J)
ForeRunner (150J)
16
14
12
Current (amps)
10
8
6
4
2
0
-2
-4
-6
-8
-10
-12
-2
0
2
4
6
8
10
12
14
16
Time (msec)
Maximum shock into 125 ohms
Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors.
For training purposes only
18
20
22
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Current Duration
25
20
Shock comparison at maximum energy level
High impedance condition (125 ohms)
Medtronic Physio-Control
(360J)
Current (amps)
15
Company B (200J)
10
Increased Duration for
Increased Impedance
5
0
0
2
4
6
8
10
12
14
16
18
20
All waveforms
generated using
actual devices
-5
-10
-15
Time (msec)
High impedance patients receive more energy with the
Medtronic Physio-Control waveform
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
ZOLL Waveform Delivers More Energy and
Higher Peak Current at 100J than MPC
100 J setting, 100 ohms
100 J setting, 75 ohms
15
5
0
-5
5
0
-5
-10
-10
-15
-15
-2
0
2
4
6
8
10
12
14
16
18
20
Physio - 102 J delivered
Zoll - 119 J delivered
10
Current (A)
10
Current (A)
15
Physio - 102 J delivered
Zoll - 120 J delivered
-2
0
2
4
6
8
10
Time (ms)
Time (ms)
For training purposes only
12
14
16
18
20
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
ZOLL Waveform Delivers More Energy and More
Current at 200J than MPC
200 J setting, 75 ohms
25
25
Physio - 203 J delivered
Zoll - 224 J delivered
20
15
15
10
10
5
0
-5
5
0
-5
-10
-10
-15
-15
-20
-20
-2
0
2
4
6
8
10
12
14
16
18
20
Physio - 205 J delivered
Zoll - 220 J delivered
20
Current (A)
Current (A)
200 J setting, 100 ohms
-2
0
2
4
Time (ms)
6
8
10
Time (ms)
For training purposes only
12
14
16
18
20
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Damage
• Clinically significant damage has not been
proven with monophasic shocks at
traditional energy levels
• Our ADAPTIV biphasic shocks give 40%
less peak current than monophasic shocks
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Damage:
None at Levels up to 360J Monophasic
Cardiac troponin I levels after monophasic shocks13
Not Elevated
Minimally elevated
(0.8–1.5 ng/ml)*
* Permanent cardiac injury occurs
at levels greater than 7.7ng/ml
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
1st Shock Comparison, Average Impedance
25
LIFEPAK 12 biphasic device (200J)
FirstSave (low)
M Series (120J)
ForeRunner (150J)
20
Current (amps)
15
10
5
0
-5
-10
-15
-2
0
2
4
6
8
10
Time (msec)
First shock into 75 ohms
Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors.
For training purposes only
12
14
16
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
The Illusion of Damage
18
LIFEPAK 12 biphasic device (360J)
FirstSave (high)
M Series (200J)
ForeRunner (150J)
16
14
12
Current (amps)
10
8
6
4
2
0
-2
-4
-6
-8
-10
-12
-2
0
2
4
6
8
10
12
14
16
Time (msec)
Maximum shock into 125 ohms
Source: Medtronic Physio-Control; Shocks delivered from commercial device into resistors.
For training purposes only
18
20
22
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Summary: Damage
• Peak current, not energy causes damage
• No documented clinically significant
damage with monophasic waveforms
• ADAPTIV biphasic waveform at 360J
offers peak current levels similar to
monophasic at 100J
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Summary: Heartstream
• Heartstream
– Extends waveform duration
– Energy remains fixed at 150 J
• Gives most current to low impedance patients
• Gives least current to high impedance patients
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Summary: ZOLL
• ZOLL
– Does good job of delivering higher level of current to
high impedance patients
– Does not ever extend duration
– Must go to BTE waveform at 200J and above 85 ohms
• But does not extend duration
– Tries to pass off Heartstream’s studies as their own
– U. Mass AF study will be discussed Thursday at lunch
time session
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
The Guidelines
• ILCOR
– What is their position?
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
International Guidelines 2000 Regarding
1
Biphasic Defibrillation
“Current research confirms that biphasic shock
energies of  200J are safe and effective.” (Class
IIa) P I-90
“Research has not yet determined the optimal
biphasic waveform.” P I-90
American Heart Association/International Liaison
Committee on Resuscitation, pp I-90–I-91.
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
The Guidelines on
Biphasic Energy Levels
“We cannot make a definitive recommendation for
the energy for first and subsequent nonescalating
biphasic defibrillation attempts.” P I-91
• “Even though both escalating or nonescalating
biphasic defibrillators are available, there was
insufficient data to recommend one over the
other.”
Source: Guidelines 2000 for Cardiovascular Resuscitation and Emergency
Cardiovascular Care.Circulation. 2000;102(suppl I) 8. August 22, 2000
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Biphasic
350
300
250
200
Monophasic
150
Biphasic
100
50
0
Spontaneous VF
Electrically Induced
Spontaneous VF
Electrically Induced VF
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Author’s Conclusions
“These results suggest that clinical studies that involve
defibrillating spontaneous arrhythmias in either a prehospital or
intensive care unit setting are necessary to determine optimal
defibrillation energy levels for defibrillation using BTE
waveforms and should include shock strengths to 360 J.”
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Biphasic Defibrillation-MDT 1st OOH
Results—46 VF patients
• MDS in LIFEPAK 12
– 28 VF patients
– Collapse to shock =
13.03 minutes
– First shock success
19/28 (68%)
– ROSC 15/28 (54%)
– Survival 5/28 (18%)
• BTE in LIFEPAK 12
– 18 VF patients
– Collapse to shock =
12.25 minutes
– First shock success
15/18 (83%)
– ROSC 15/18 (83%)
– Survival 8/18 (44%)
Doets, vanAlem, Koster presentation at the World Congress in Australia, May 2002
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Reality
• LIFEPAKs have
– Current Duration
– Voltage Compensation
PLUS
– Optional energy settings
• Lower risk of refibrillation
• Higher second shock success
• Wider range of therapy
• Trend toward more survivors
– Goal: to save more lives
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Predictions
• More studies will show biphasic is better
in OOH setting
• Customers will become more interested
(and confused) in biphasic defibrillation
For training purposes only
Technical
EuropeanService
Sales Meeting
Training
Barcelona 2002
Predictions
• More manufacturers will try to build devices
capable of >200J
• The Guidelines will remain open to
interpretation
• More lives will be saved
For training purposes only
Download