My name is Maddie, I’m a Fort Smith tech, member of the QA team, and we’re going to go over what will be a Cadwell refresher for most of you today, along with an introduction to a new modality for most of you. Channel Setup • -Step One-Input Editor This is where your individual channels live, or where you plug in individual needles. You have four pods, two amplifiers. You cannot have more than sixteen channels on each amplifier ENABLED for the system that most of you have. What this means is that if you have more than 16 channels on the mode group you have selected (SSEP, EMG, etc.) you will have to go and deselect them. I’ll touch more on this later. The number of individual inputs isn’t necessarily important. • The only real consideration with what goes in what “pod” is accessibility to the patient. You wouldn’t want your head channels on a pod running down to the foot. Input Editor Step Two • Within whatever pod and amplifier you’re looking at, click the small drop down icon, and begin typing what you’re looking for. Functionally, it does not matter what this is called, just the laterality. The label can be changed to whatever you want later. It’s just to keep you from confusing yourself • These are paired, and will normally pair themselves together in active/reference. If not, just search for the same thing and add the opposite polarity ( (-) (+) ) below your first channel. Channel Editor- Adding Input to Channel • Once you’ve created your inputs, go to the right side and find your amplifier • Click the drop down under “active” and find one of your inputs. Then, go to reference, click the drop down, and find your other input. • Next to that is the name—you can name your channel whatever you want. Remember to add innervations Channel Editor and Mode Group Errors Active and reference must be from the same amplifier- they can be from different pods, but not one from amp one, and the other from amp two I’ll touch on this later. If you’re creating a test (this red banner suggests that you are) this is a real error and you’ll have to disable these channels, or deselect the mode that they are assigned to Mode EMG must be included in a test to be included in Mode Group EMG and EEG- you’ve assigned EMG to a mode, but haven’t included the mode group in an active test, so it’s throwing this. Seems silly, but that’s what it is Filters, Settings: Gain • This is not the same type of display gain that you engage with when you scroll up or down on an existing waveform. Input editor gain controls how large or small that the waveform has the capability to display or record at. • If the waveform is too large or is being clipped, you can increase the display gain. If it’s too small, you can lower this. • For SSEPs, usually starts at 10uV/div • For EMG/MEP, usually starts at 500uV/div • For To4, usually starts at 5000uV/div To4 Gain Any time you see that boxy morphology on To4 or MEP (it can happen in fEMG too, but usually with EKG artifact) open up your gain, and it will usually resolve. It’s a bigger deal with MEPs as far as interpretation, but it certainly helps here, since the amplitude doubles. I’d say it started at 5,000uV and then increased to 10,000uV Filters, Settings: HFF/LFF • Whether it’s high or low cut (frequency) filter, the purpose of these are to attenuate frequencies that are above or below your desired frequency • High frequency will attenuate frequencies over what you have it set at. Low frequency will attenuate frequencies below what you have it set at • SSEP- HFF=250, LFF=30 • EMG/MEP- HFF=2500, LFF=30 • EEG- HFF=70, LFF=1 • To4- HFF=1000, LFF=10 • TLDR- These are set at frequencies that you wouldn’t *typically* see traces averaging with frequencies above or below. • Raise the HFF if you want something higher to come through, lower the HFF for the same reason. Filters, Settings: Notch Notch is a combination high and low frequency filter designed at a band to filter out 60hz electrical noise (in the US) and decreases power for the band between 50-60hz It can be a good tool to have, but it’s relatively controversial, and most neurologists prefer not to use it, since it can attenuate both the electrical noise and evoked/free run potentials Notch affected the final amplitude and latency of evoked potentials, filtering can result in considerable distortion of the time course and amplitude of a signal demonstrated It can also cause ringing artifact in SSEP, tEMG, and some cortical attenuation in SSEPs, although I believe that’s a little less common This is another example of distortion, but in tEMG Mode Controls: SSEP Sweep-How much data is actually being recorded Amp Channel/ESU- leave off, can cause rejection Auto Store- DON’T turn off Rep rate- not divisible by 6, lower the rep rate, typically better the stimulation Filter- higher the filter number, the more distortion you’re getting Invert-If your waveform looks inverted, please fix it . It’s not an aesthetic issue, the waveform will degrade over time, and the latency can appear to be falsely prolonged or the amplitude decreased if it’s not the appropriate polarity Enable- Just whether or not you want it on the screen or not. You can also reach this from the “gear,” but it will still trace. This is turning off the trace completely Reject- This is for specific channels (like C5, EP, PF) that have a lot of artifact. You can just turn that channel’s reject off and leave the cortical channels unaffected Mode Controls: MEP Sweep: Tall, myelopathic etc. patients, prolong this to 12-15 for single train if you don’t see an AH response Auto lock: don’t turn off Auto lock duration: how long it remains on Primary stimulator: Can change the output if you’re doing corticobulbar MEP or if your output is defective Polarity: If you can’t get to the head and you’re only seeing contralateral, reverse this MEP double train • Pulse count- I like 2 and 7, like Peyton said, 2+7 does not equal 9 This seems to be the “happy number” of priming vs. stimulating • ITI- Intertrain interval. Needs to be below 20 or above 100 • (?) for optimal stimulation. This separates priming and stimulation. I like 15ms • ISI- Interstimulus interval. The amount of time between grouped pulses. 3.5 tends to be happy for LE, 1.5 tends to be happy for UE, 1.0 or so tends to be happy for face. Play around and see what works • Keep polarity the same on both • Sweep needs to be at least 15 ms/div if you’re doing 15 ms ITI. Proximal UE muscles can bleed into the stim artifact Mode controls tEMG Reject/amp channel- This is really what I want to highlight here. If you’re doing a thyroid or any other stimulation case, uncheck these, and the “high impedance trials,” it will allow the system to process every input I would also suggest, as we said earlier, not leaving the notch filter on channels that are applied to tEMG to avoid that distortion Cascade/cadwell Errors: • These errors can usually be divided into one of two situations: you have a physical problem with your amplifier (or surrounding components), or Cascade has become overwhelmed by the demands of life • No base module connected- sometimes this literally means no base module connected. Other times the power supply isn’t working, the plug-in to power supply is too loose/failing, the aux cable has come loose/is failing, or it’s a software failure. It usually cycles through this if it’s a software failure. • Amplifier can only support 16 channels- 99% of the time, this means that one of your physical amplifier connectors isn’t seated well. Unscrew the (tiny) screws, push everything in to ensure a good connection, screw the (tiny) screws back in. • Stimulator not connected- either a disconnect of the stimulator cable on the box end, the amplifier end, or the program is cycling through “base module” physical errors before it starts • The initial program itself will not launch- literally turn it off or turn it back on and check aux cable/power supply. Make sure your software is compatible with the amplifier you’re using, if the amp is further updated than the computer, it will not launch Cascade/Program Errors- The operation could not be performed while the system loaded configuration into hardware- most of the time, this is when the program itself has buffered or frozen, and is cycling back. It should go away on it’s own ES-IX IBlock warning- amplifier saturation or electrical interference with stimulation, run impedance to desaturate the amp Mode failed to initialize without reason- this typically follows the ES-IX IBlock warning when you run impedance to desaturate the amp, but cautery resumes while impedance is running. It should go away on it’s own and is something that is typically cycled through Stimulator output level is low, patient connection may be bad- this is actually helpful, most of the time. It’s an imbalance in requested versus delivered voltage. However, if you’re getting this error and your connection is fine, it’s either the port that’s gone out, or potentially the metallic portion inside of the stimulating leads have gotten bent/broken and can no longer stimulate effectively. I have also had this happen when the system is both giving and receiving stimulus information. Common Forms of Artifact: What does this look like to anyone? It’s amplifier saturation, but what makes this a little bit unique is there’s still a couple of channels in EEG and SSEP that are acquiring noise that looks somewhat like a waveform, so that can be confused with an issue without context What does this look like to anyone? It could be one of a few things, but the defining feature here is that the one channel that is clean is Cz-Fz, so you’re not getting as much noise from the temporalis, frontalis, masseter in that channel as you are the channels that aren’t referenced to Cz. This requires context clues as well, if the patient has no myogenic artifact in any other modality, no blood pressure increase, is isoelectric, has been stable throughout the case, you’re probably looking at an electrical issue What does this look like to anyone? Noise, someone touching wires, poor impedance? It’s actually nothing plugged in at all. The pod is still connected to the amplifier, but there’s no needle plugged in. It looks scarily like real data, so always go visually verify that your needles are still in What does this look like to anyone? For context, this was a thoracic lam with kypho, I think for malignancy. With that history, I thought this was real, but as you can see, it’s very cyclical and too rhythmic. Given the channels and location, where do you think it is? It’s a gastric stimulator. I had no medical history on this patient, but eventually figured this out and confirmed this with a merged medical record. Let’s talk about transabdominal MEPs quickly TaMEPs are elicited by stimulating over the conus medullaris, and thereby, the ventral nerve roots, directly TcMEP activate reduced subset of lower motor neurons whereas TaMEP activate nearly all spinal nerve fibers (Allison et. al., 2024). There has been some concern about how you know that you are activating what you think that you are, that’s slightly above my pay grade, but I would venture to say that you are getting some current spread from what I’ve seen Relatively resistant to anesthesia, think D-wave (there are differences in terms of averaging, why/how/where you run them, but they are both nonsynaptic) whereas TcMEPs are sensitive due to number of synapses ONLY pertains to lumbar, L2 or below; no cervicothoracic applications Should be used in conjunction with TcMEPs as a form of control DIY: TAMEP Anode-single large ground sticky pad at umbilicus Cathode-two needle electrodes “directly caudal to the tip of the conus medullaris,” at L12, straddling spinous process 1 pulse, ~300V, 3.5ISI Use a separate cord and output, reserving one for TcMEP TAMEP Execution Notes: Need fluoro for these, especially on higher BMI patients; can palpate from T12 rib and count spinous processes, but easier to fluoro. I ended up slightly more distal (not markedly so where my results were impacted or where I would just be getting current spread) and slightly eccentric to one side Positive correlation with voltage and BMI (Allison et. al., 2024). Higher BMI, higher stimulation Use Tegaderm for anterior and posterior leads, prep over posterior leads. For posterior fusions (not lateral) the surgeon you’re working with will have to be okay with these in the prep field Can reverse polarity if you’re not getting response you think you should be These could hypothetically move if they expose and manipulate the fascia around it (I had this concern, luckily didn’t happen) I think it is of more concern for large open fusions vs MIS exposures These would be much more helpful for something like a lateral paired with saphenous SSEP Be mindful of movement; causes kind of “dolphin jump”, have surgeon remove hands Make sure you’re using them in indicated procedures as to not supersede MEP consent. You wouldn’t want to run this on anything it’s not specifically indicated for, or it would be considered more experimental and unnecessary