Technical Service Manual StarLux500 StarLux500™ PULSED LIGHT AND LASER SYSTEM TECHNICAL SERVICE MANUAL Document Part Number 5500-0000 Revision X1 Manufactured by: Palomar Medical 82 Cambridge Street Burlington MA 01803 USA Toll Free 800-725-6627 Tel. 781-993-2315 Fax 781-993-2340 Email: customer_service@palomarmedical.com www.palomarmedical.com Header Page P/N 1500-0018 rev X1 Technical Service Manual StarLux500 Chapter 1 Safety 1.1 Safety Scope 1-2 1.2 Training 1-2 1.3 Safety Overview 1-3 1.4 General Safety 1-4 1.5 Room Safety 1.5.1 Electrical Safety 1.5.2 Fire Safety 1-4 1-5 1-5 Chapter 1 P/N 5500-0000 rev X2 1-1 Technical Service Manual StarLux500 1.1 Safety Scope This manual is designed to help technically qualified personnel of Palomar Medical Products and authorized distributors to repair and maintain the StarLux™500 Pulsed Light and Laser System. This manual contains functional descriptions of the StarLux500, as well as diagrams, schematics and parts lists. This information, which Palomar considers proprietary, is not intended for use by, or disclosure to, anyone not associated with Palomar Medical Products. 1.2 Training Service personnel should: a) Have a thorough understanding of how the StarLux500 is used for treatments. b) Be completely familiar with the Operator's Manual provided with the StarLux500 System. c) Have knowledge of analog and digital electronics. Extent of Service: Troubleshooting should normally be to the subassembly level. Groups of components, rather than individual components, should be replaced in order to retain factory adjustment of functional subsystems. Chapter 1 P/N 5500-0000 rev X2 1-2 Technical Service Manual StarLux500 1.3 Safety Overview The Palomar StarLux500 is an electro-optical device intended for safe and reliable hair reduction, pigmented lesion treatment, leg vein treatment, acne treatment, deep tissue heating, and soft tissue coagulation. Personnel operating and maintaining the StarLux500 should be familiar with the safety information provided in this section. Palomar assumes no liability whatsoever for any damage or injury resulting from an application of a product which is not in strict accordance with the instructions provided with the product. Palomar also assumes no liability for any damage or injury arising as a result of user error or mistake, including but not limited to injury arising from user’s lack of qualification to administer the treatment in the individual circumstances, or as a result of errors or mistakes committed by such user who would otherwise be considered as having the necessary professional skills to apply such treatment. Read the Installation and Operation Instructions thoroughly before connecting the base system and chiller unit to the main power connection prior to use. The system must be set, regulated and used in accordance with the instructions for use listed in the handpiece appendices. Failure to observe usual safety precautions may present a risk of hazardous exposure to laser or pulse light radiation. All persons operating the StarLux500 System or in the vicinity of the system must be aware of the potential hazards of the system output and should take the proper safeguards as described in the Operator’s Manual. Be certain all personnel carefully review the safety precautions pertaining to the Palomar StarLux500 System. Only authorized individuals with appropriate safety training and knowledge should operate, assist in the operation of, or provide maintenance/service to the system. One person should be assigned to be responsible for system controls during the procedure. High voltages are present inside the StarLux500 System, and any pulsed-light or laser emitting device can cause injury if used improperly. Personnel who work with pulsed light and laser sources should always be aware of the possible dangers and should take the proper safeguards as described in the operator’s manual. The Palomar StarLux500 System is an “eye-safe” device when operated properly, however, as with any laser and pulsed light device, never look directly at the light output. During treatments, all persons must wear protective eyewear to protect against possible eye damage. No one other than Palomar Medical authorized personnel may service inside the protective covers of the StarLux500 System. Dangerous voltages are present inside the system. Maintenance should be performed by the operator 30 minutes after the system is shut down and disconnected from power. Performing maintenance procedures with the system while it is powered up or before 30 minutes from disconnecting the system, may be hazardous to the operator and destructive to the system. Chapter 1 P/N 5500-0000 rev X2 1-3 Technical Service Manual StarLux500 1.4 General Safety • Delivering excessive energy to the treatment site may cause thermal damage to the skin, resulting in superficial damage to the skin, crusting, skin redness, hypopigmentation or hyperpigmentation. • Consumer safety is assured by a well-trained staff. • Consumer education is also important and information about the nature of the treatment should be provided. • Never look directly at the light coming from the distal end of the handpiece as this may cause eye injury. • Never direct the laser or pulsed light at anything other than the targeted area. • Never point the handpiece so that it discharges into free space. Make sure it is pointed at, and in full contact with, the treatment site during treatment or when performing test spots. Warning Do not submerse the handpiece in any solution. Submersing the handpiece may cause a hazardous condition for the user and possibly the consumer. Submersing may also cause the system to malfunction. Apply disinfecting or cleaning solution to wipes or gauze pads to be used on the handpiece. If the handpiece becomes submersed in fluid or leaks any cooling water, discontinue use of the handpiece immediately and call Palomar for replacement. Do not use the handpiece if it is cracked, the sapphire lens missing, cracked, or burned. 1.5 Room Safety • Identify the treatment room clearly. • Restrict entry to the treatment room when the system is in use. Allow access to the treatment room only to those personnel both well trained and essential to the procedure. • When the system is not in use, ensure that it is rendered inaccessible to unauthorized personnel. Remove the key to disable the unit. • Make sure that all personnel are familiar with the system controls and know how to shut down the system instantly. • Block all appropriate windows, etc. to prevent light from escaping the treatment room. 1.5.1 Electrical Safety Chapter 1 P/N 5500-0000 rev X2 1-4 Technical Service Manual StarLux500 • The electrical safety hazard of any laser and pulsed light system cannot be ignored or taken lightly. Although the Palomar StarLux500 System has many safety features in it’s design, certain precautions must be observed. • It is possible for the high voltage components to retain a charge even after the system has been turned off. Therefore, no part of the exterior housing of the main system or handpiece should be removed except by trained and authorized technicians. • This may expose personnel to dangerous high voltages and potentially dangerous light radiation. Only a qualified Palomar Service Representative must perform any service requiring access to the interior of the instrument. To contact a Service Representative, dial toll free 1-800-PALOMAR or 1-781-993-2315. • Customer Service Email: customer_service@palomarmedical.com 1.5.2 FIRE SAFETY The potential for fire hazard exists because of the nature of light treatment. The absorption of emitted energy, no matter how brief, may raise the temperature of any material. This phenomenon is the basis of many useful medical and surgical applications; it is also the reason that these applications often require precautions against the risk of igniting combustible materials in and around the treatment area. When the Palomar StarLux500 System is used, the following precautions should be taken: • Anesthetics administered must be approved as non-flammable. • Ensure that the system and ancillary equipment are working properly before beginning a procedure. • Never direct the light output at anything other than the intended treatment site, especially at reflective objects such as metal surgical instruments. • Be certain the footswitch is clean and working properly. Place the footswitch where it will not be mistaken for another piece of equipment. • Exercise particular care in the use of oxygen. The presence of oxygen can accelerate the combustion of any flammable material. • Avoid using combustible materials, such as gauze and drapes, in the treatment area. When they are required, these materials may be made fire-retardant by keeping them moist with water or saline. Clothing should be kept well away from the area of treatment. • Know the location and operation of the nearest fire extinguisher. • Always turn off the system and remove the key when ever possible. This will prevent accidental light emissions. Chapter 1 P/N 5500-0000 rev X2 1-5 Technical Service Manual StarLux500 Chapter 2 Overview 2.1 Introduction 2-2 2.2 System Specifications 2-3 2.3 Theory of Operation 2.3.1 General Description 2.3.2 Sequence of Operation 2-4 2-4 2-4 Chapter 2 P/N 5500-0000 rev X2 2-1 Technical Service Manual StarLux500 2.1 Introduction The Palomar StarLux™500 Pulsed Light and Laser System is designed, by attaching multiple treatment handpieces, to deliver pulses of broadband incoherent light or laser light to a predetermined target site. The Palomar StarLux500 System is specially developed for permanent hair reduction, pigmented and vascular lesion treatment, leg vein treatment, acne, fractional soft tissue coagulation, fractional skin tightening through soft tissue coagulation, and other clinical applications. The Palomar StarLux500 System is user friendly, allowing the treatment provider to select output fluence (J/cm2) and pulsewidth (ms). This enables the user to safely perform treatments on skin types I-VI as indicated in Starlux Treatment Guide. The system is housed in a protective cabinet with includes the connection to an external cooling unit. Delivery of the energy to the treatment site is through a handpiece that contains a cooled sapphire window. Light is directly administered to the treatment area by placing the handpiece window in contact with the skin and pressing an attached footswitch in conjunction with pressing the handpiece trigger button. There is also an option to deliver light pulses by pressing the footswitch only. The “Cooltip” supplies contact cooling to the skin during treatments, this minimizes the potential for skin damage and allows for higher fluence during treatment. Chapter 2 P/N 5500-0000 rev X2 2-2 Technical Service Manual StarLux500 2.2 Palomar StarLux500 System Specifications Item Description System Type Class I Equipment, Type BF, applied part (CE Approval) Class II Medical Device (USA) Pulsed Light Class IIb (EC Community) Class IV Laser (Lux1064TM and Lux1540 TM Fractional Handpiece only-USA) Wavelength 400nm - 1540nm (Handpiece Specific) Spot Size Handpiece Specific Pulsewidth Handpiece Specific Fluence Handpiece Specific Light Delivery Flashlamp or Laser within the handpiece (Handpiece Dependent) Repetition Rate Handpiece Specific Actuator Footswitch and/or Handpiece trigger Voltage Max Power Input 100 - 240VAC, 50/60Hz 1000VA (Base Unit) 650VA (Cooling Unit) 1650VA (Total System) Operating Environment Temperature: 15-28° C (55-80° F) Humidity: 0 to 40% Transport/Storage Temperature: 0-50° C (32-124° F) Environment Humidity: Below 80% System Cooling Cooling system for flashlamp Cooling system for sapphire Dimensions Base Unit: 19” (48 cm) L x 11” (29 cm) H x 14” (36 cm) D Cooling Unit: 10 “ (27 cm) L x 15” (38 cm) H x 16” (41 cm) D 500 Cart: 22 “ (58 cm) L x 30 “ (78 cm) H x 22 “ (58 cm) D Weight Base Unit: 55 lbs. (25kg) Cooling Unit: 33 lbs. (15 kg) 500 Cart: 55 lbs. (25 kg) Chapter 2 P/N 5500-0000 rev X2 2-3 Technical Service Manual StarLux500 2.3 Theory of Operation 2.3.1 General Description: The StarLux500 system uses Analog and Digital Logic Circuits, a Microprocessor and Software, High Voltage, and Flashlamp or Laser to create pulses of light for treatment. The system can operate with input voltage range 85VAC to 265VAC 50Hz/60Hz. The system’s internal power supply adjusts to the incoming line voltage. The system contains sub-assemblies as described below: 1. External Cooling Unit, Provides cooling for the CoolTip in the handpiece, monitors the water flow and temperature of the lamp and Cooltip loops. 2. TEC module (3 Dual Chamber TEC’s) within the cooling unit generates cooling for the Cooltip. 3. Display Module, Contains the Main Processor, interconnect board and display screen. 4. Power Supply Module, provides the DC voltage to operate the System, Lamp Controller and Thermocontroller boards. 5. Handpiece, provides the light to the target tissue and contact cooling. All calibration data stored in handpiece. 2.3.2 Sequence of Operation: 2.3.2.1 A power cord is connected to both the base unit and external cooling unit, a handpiece is connected and in the holster, and the coolant reservoir is full. Turn the main power switch on the cooling unit and then the base unit. AC voltage is then applied to the Power Factor Controller Board. This converts the base unit input voltage to supply 390VDC to the capacitor charging circuit and Thermocontroller board. This board supplies 12VDC and 6VDC to all boards and the display module. A “Menu/Next” screen will be displayed. (screen 1) Screen 1 Screen 2 The system will read the EEprom information from the handpiece, the software will determine the type of handpiece installed and then load the proper parameters for the handpiece. (screen 2) If a handpiece communication error is detected during this time, it will be displayed on the screen. Otherwise, the system will proceed to key switch screen. Chapter 2 P/N 5500-0000 rev X1 2-4 Technical Service Manual StarLux500 2.3.2.2 With the key switch in the off position, the ‘turn key’ symbol will be displayed on the touch screen panel. Note that the colored band at the top of the screen is the same color as the handpiece installed. (screen 3). Screen 3 Screen 4 2.3.2.3 When the key is turned to the “ON” (clockwise) position, the microprocessor will check all operating systems for voltages or other faults and the cooling module power will then switch on. As the system continues to power up, checks for water flow and temperature will also take place. The lamp will start simmering. (screen 4) Screen 5 (Cool Down) 2.3.2.4 If the system does not have any immediate faults, the “Cool Down” screen with the hourglass symbol will be displayed.(screen 5) The cooling module is now fully operational and if there are any faults not recognized during the system startup test, they will be displayed now. The coolant reservoir holds about 350cc of cooling fluid. It is 2 chamber tank separated for the Cooltip loop and the Lamp loop. Lamp Loop Cooling: If the lamp loop temperature is 35˚C or higher, the fans run at 100% speed. If the lamp loop is under 30˚C, then the fans will run at 80% speed. Two Thermisters (Lamp & Cooltip) monitor the water temperature and must be within 4˚C of each other. Lamp Water flow is >1.9 LPM. The Pressure switch is rated at 65 PSI and will open if the pressure rises above this point. Chapter 2 P/N 5500-0000 rev X1 2-5 Technical Service Manual StarLux500 CoolTip Cooling: A TEC module provides the cooltip cooling. Cooltip Water flow is >0.45 LPM. The Pressure switch is rated at 65 PSI with a bypass circuit to prevent damage if water flow is blocked. The “hourglass” may be displayed above the thermometer symbol on the left side of the display screen. During this mode, the cooltip temperature is being monitored while the TEC module is cooling the water and the hourglass symbol will stay on until the water temperature is 15˚C (59˚F). This should occur within 7 minutes. If the room temperature is above 22˚C (72˚F) or the system has been operating for a long period of time and then shut off, it may take longer for the cooltip temperature to reach 15˚C. The system will not go into the “Ready” mode until the hourglass is gone from the screen. (screen 5) The Starlux must be in Ready mode in order for a handpiece to pulse. Screen 5 (Ready Mode) 2.3.2.5 From the “Ready” screen, the user can select pulse width (ms), fluence (J/cm2), and repetition rate (Hz) up to the maximum parameters for the handpiece. For questions about handpiece or system information (ie. hp shot count or system software version), select the question mark button and choose Handpiece or System. (screen 6 and 7) Select ‘Back’ button to return to your previous screen or to exit back to the Ready screen. Screen 6 Screen 7 (System Information) 2.3.2.6 For complete system operation please see Chapter 4 of StarLux500 Operator’s Manual. Chapter 2 P/N 5500-0000 rev X1 2-6 Technical Service Manual StarLux500 Chapter 3 3.1 Chapter 3 P/N 5500-0000 rev X2 Troubleshooting Fault Codes 3.1.1 Non Error Code Faults 3-2 3-14 3-1 Technical Service Manual StarLux500 3.1 Fault Codes Verifer StarLux System Faults Error Code Possible Cause Customer Correction 1. 2. 001- 1 Internal Voltage exceeded limits. 3. Restart system. If problem still exists, try another handpiece. Call Palomar Service. 1. 001001- 2 3 Internal Voltage exceeded limits. Not used. Restart system. If problem still exists, try another handpiece. 2. Turn off power switch at the back of the system. Let system cool off for 5 minutes. 3. Restart system. 4. Call Palomar service. Not used. 1. 2. 001- 001- 001- 001- Restart system. Ensure that air intake and exhaust are not obstructed. o Cool room down below 80 F. Let system cool down. Call Palomar service. Shut down system and reconnect handpiece Try another handpiece. 1. Check wiring on handpiece. Call Palomar service 2. Check wiring J4 interface board. Install Remote Interlock Connector in back of the system. 1. Check Remote Interlock harness Call Palomar Service. and connector. 4 System Cabinet temperature exceeds allowable limit. 5 Handpiece is disconnected from the system. 6 Remote Interlock Connector not plugged in. 2. Emergency stop button is pushed in. 1. Turn off the power switch at the rear of the system. 2. Release Emergency stop button by turning red knob clockwise in same direction as arrow. 3. Restart system. 4. Call Palomar service. 7 Chapter 3 P/N 5500-0000 rev X2 3. 4. 1. 2. 3. 1. Detailed Troubleshooting 1. Check +12vdc at J3 interconnect board. (11.4 to 12.6vdc.) If not in range, replace Large Condor (GSM28-12) power supply on ThermoController bd assy. 2. If voltage ok, replace display module. 1. Check +5vdc from Voltage Regulator Bd Assy to System Interconnect Bd. Assy. J4 pin 9. 2. If +5 is supplied to Display Module, let system sit in keyswitch mode while monitoring +5v. 3. If the +5v shuts off, call Palomar service. Not Used. 1. Using System Debug Software, Check Cabinet Temperature Sensor for correct operation. 2. If temperature is wrong, ensure that the System Interconnect board is properly mounted and the connector is fully connected to the Interface board. 1. Check EMO harness and connector. 2. Check EMO switch. Sw Closed = System On. (button pulled out) Sw Open = System Off. (button pushed in) 3. Verify board revision. 3-2 Technical Service Manual StarLux500 1. Restart system. 2. Try another handpiece. 3. Call Palomar service. 001- 8 Internal Voltage loaded down. 001- 9 Water leak in handpiece. 1. Try another handpiece. 2. Call Palomar service. 001- 10 001- 11 1. 2. 1. 2. 001- 12 001- 13 14 Wiring came loose in handpiece. Internal Voltage loaded down. A Critical error occurred within the PFC power supply. A Critical error occurred within the ThermoController power supply. System Handpiece Cooltip thermal sensor disconnected or temperature between two thermistors greater o than 4 C. 1. Let the system cool down. 2. Restart the system. 3. Call Palomar service. 15 System Flashlamp thermal sensor disconnected or temperature between two thermistors o greater than 4 C. 1. Let the system cool down. 2. Restart the system. 3. Call Palomar service. 001- 001- 001- 16 001001- 17 18 Internal ThermoController driver circuitry has excessive current. Internal ThermoController driver circuitry has excessive current. Not used. 001- 19 Internal Voltage loaded down. Chapter 3 P/N 5500-0000 rev X2 Try another handpiece. Call Palomar service. Try another handpiece. Call Palomar service. 1. Verify +12vdc is supplied from Small Condor power supply (GSM7-12) on ThermoController bd. 2. Replace power supply. 3. Replace display module. 4. Verify all cable connections. 1. Check Hand piece for water leak or condensation. 2. Check Hand piece for shorted water leak sensor. 3. Replace handpiece. 1. Check Umbilical wiring. 2. Check handpiece board connectors. 3. Replace handpiece. 1. Replace handpiece. 1. Restart the system. 2. Call Palomar service. 1. Replace power supply module. 2. Call Palomar service. 1. Restart the system. 2. Call Palomar service. 1. Replace power supply module. 2. Call Palomar service. 1. Monitor thermistor temperature readings. Temperatures should o be within 4 C of each other. 2. Check connection to Handpiece Loop Thermistor. Verifiy Thermistors are good and at same values. 1. Monitor thermistor temperature readings. Temperatures should o be within 4 C of each other. 2. Check connection to Flashlamp Loop Thermistor. Verify thermistors are good and at same values. 1. Check Fan wiring and connector. 2. Ensure Fans are free moving. 3. Replace fan assembly. 4. Check ThermoController board. 5. Replace power supply module. 1. Check TEC resistance. 2. Check TEC wiring. 3. Check ThermoController Board. 4. Replace TEC module. Not Used. 1. Measure +24vdc on ThermoController board. 2. Replace power supply module. 1. Restart the system. 2. Call Palomar service. 1. Restart the system. 2. Call Palomar service. Not used. 1. Restart the system. 2. Call Palomar service. 3-3 Technical Service Manual StarLux500 20 21 Internal Lamp Cooler temperature is o beyond 50 C. Not used. 001- 22 Handpiece cooltip water flow is restricted/low. 001- 23 Water Level is too low. 001- 24 Flashlamp water flow is restricted/low. 001001- 001- 25 001- 26 001- 27 Internal Chiller temperature is o beyond 50 C. A Critical error occurred within the LampController power supply. Lamp Current exceeds limits. Chapter 3 P/N 5500-0000 rev X2 1. Let the room cool down o below 80 F. 2. Ensure that the system has adequate space for air flow. Make sure that there 1. Check Lamp temperature with are no air restrictions or Debug Software. blocked vents Let the 2. Check Heat Exchanger Fans system cool down. operation. 3. Check water flow rate. 3. Restart the system. Let 4. Check flashlamp water sensor. system run without emitting light for 2 minutes. 5. Check ThermoController. 6. Replace power supply module. 4. Call Palomar service. Not used. Not Used. 1. Verify rear fluid bypass jumpers are connected 1. Check handpiece water flow properly. rate. 2. Fill reservoir with Palomar 2. Check handpiece flow switch. fluid. Sw Open = No Flow 3. Ensure handpiece is properly seated. Sw Closed = Flow 4. Change water filters. 3. Replace handpiece flow switch. 4. Check wiring. 5. Try another hand piece. 5. Replace Cooling Module. 6. Call Palomar service. 1. Check reservoir level verify filled with water. 1. Fill reservoir with Palomar 2. Check for leaks. fluid. 3. Check water level switch. 2. Call Palomar service. 1. Verify rear fluid bypass jumpers are connected 1. Check flashlamp water flow properly. rate. . 2. Fill reservoir with Palomar 2. Check flashlamp flow switch. fluid. 3. Ensure handpiece is Sw Open = No Flow properly seated. Sw Closed = Flow 3. Replace flashlamp flow switch. 4. Change water filters. 4. Check wiring. 5. Try another handpiece. 5. Replace Cooling Module. 6. Call Palomar service. 1. Check HP temperature with Debugger Software. 1. Ensure that the system 2. Check Chiller cooling has adequate space for air performance. flow. Make sure that there 3. Check Heat Exchanger Fans are no air restrictions or operation. blocked vents 4. Check water flow rate. 2. Let the system cool down. 5. Check handpiece water 3. Restart the system. Let sensor. system run without 6. Check ThermoController. emitting light for 2 minutes. 7. Replace cooling module. 4. Call Palomar service. 1. Restart the system. 2. Call Palomar service. 1. Replace power supply module 2. Call Palomar service. 1. Restart the system. 2. Try different operating set points. 3. Call Palomar service. 1. Verify LampController has latest revision. 2. Replace power supply module. 3. Call Palomar service. 3-4 Technical Service Manual StarLux500 001- 28 29 30 31 32 33 34 35 36 37 38 39 40 41 Lamp Energy exceeds limits. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. 001- 42 Internal PFC Operating State error. 001001001001001001001001001001001001001- 001- 43 001- 44 Verify AC input is within 90Vac to 250Vac. Internal PFC output voltages exceeds 400Vdc +/- 10v. 45 PFC temperature o exceeds 70 C. 001- 001- 001- 1. Restart the system. 2. Try different operating set points. 3. Call Palomar service. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. 1 Restart the system. 2 Call Palomar service. 1. If AC is within 90Vac to 250Vac, restart the system. 2. Call Palomar service. 46 Discharge resistor temperatures exceed o 120 C. 1. Restart the system. 2. Call Palomar service. 1. Ensure that the system has adequate space for air flow. Make sure that there are no air restrictions or blocked vents. 2. Let the system cool down. 3. Restart the system. Let system run without emitting light for 2 minutes. 4. Call Palomar service. 1. Ensure that the system has adequate space for air flow. Make sure that there are no air restrictions or blocked vents. 2. Let the system cool down. 3. Restart the system. Let system run without emitting light for 2 minutes. 4. Call Palomar service. 47 Internal Charger voltage exceeds limits. 1. Restart the system. 2. Call Palomar service. Chapter 3 P/N 5500-0000 rev X2 1. Verify LampController has latest revision. 2. Replace power supply module. 3. Call Palomar service. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. 1. Check for intermittent PFC Temperature errors. 2. Check PFC/Charger for latest revs. 3. Replace power supply module. 4. Call Palomar service. 1. Check PFC/Charger Board. 2. Replace power supply module. 1. Check PFC/Charger Board. 2. Replace power supply module. 1. Check all connections to PFC board. 2. Replace power supply module. 1. Verify that temperature sensor is steady and reading at normal voltage. If abnormal or unsteady, replace power supply module. 1. Check Charger voltage output. Voltage should be less than 410vdc. 2. Replace power supply module. 3-5 Technical Service Manual StarLux500 001- 48 001- 49 Internal Charger temperature exceeds o 70 C. Internal Charger voltage exceeds limits. 50 Internal Charger temperature sensor disconnected. 001- 001- 51 001- 52 001- 53 Internal Charger temperature sensor disconnected. Internal ThermoController Operating State error. Handpiece coolptip water pressure is high. 1. Ensure that the system has adequate space for air flow. Make sure that there are no air restrictions or blocked vents. 2. Let the system cool 3. Restart the system. Let system run without emitting light for 2 minutes down. 4. Call Palomar service. 1. Restart the system. 2. Call Palomar service. 1. If temperature is unstable or operating abnormally replace power supply module. 1. Verify PFC/Charger has latest revs. 2. Call Palomar service. 1. Let the system cool down. 2. Restart the system. 3. Call Palomar service. 1. Let the system cool down for 5 minutes. 2. Restart the system. 3. Call Palomar service. 1. If temperature is unstable or operating abnormally, replace power supply module. 1. Restart the system. 2. Call Palomar service. 1. 2. 1. 2. 3. 1. Verify rear fluid bypass jumpers are connected properly. 2. Ensure handpiece is properly seated. 3. Change water filters. 4. Try another handpiece. 1. If temperature is unstable or operating abnormally, replace power supply module. 4. 5. 1. 2. 3. 4. 001- 54 001- 55 001- 56 Flashlamp water pressure is high. Internal ThermoController driver circuitry has excessive current. Internal ThermoController communication error. Chapter 3 P/N 5500-0000 rev X2 1. Verify rear fluid bypass jumpers are connected properly. 2. Ensure handpiece is properly seated. 3. Change water filters. 4. Try another handpiece. 5. Call Palomar service. 5. 6. 7. 1. 2. 3. Replace power supply module. Call Palomar service. Check/replace water filters. Check for kinked hoses Check/replace handpiece loop pressure switch. Sw Closed = Pressure OK. Sw Open = Pressure too high. Verify Cooling Module is at latest revisions. Replace cooling module. Check/replace water filters. Check quick connects on heat exchanger. Check for kinked hoses. Check/replace flashlamp loop pressure switch. Sw Closed = Pressure OK. Sw Open = Pressure too high. Check for blocked water fittings. Verify Cooling module is at latest revisions. Replace cooling module. Check current to pump. Check pump wiring. Check ThermoController Board. Replace power supply module. 1. Restart the system. 2. Call Palomar service. 4. 1. Restart the system. 2. Call Palomar service. 1. Replace power supply module. 2. Call Palomar service. 3-6 Technical Service Manual StarLux500 001- 57 001- 58 001- 59 ThermoController temperature sensor disconnected. ThermoController temperature exceeds o 80 C. Internal ThermoController voltage exceeds limits. 001- 60 001- 61 001- 62 Internal voltage to ThermoController exceeds limits. Internal LampController Operating State error. Internal LampController Communication error. 001- 63 Simmer Error. 001- 64 1. Ensure that the system has adequate space for air flow. Make sure that there are no air restrictions or blocked vents. 2. Let the system cool down. 3. Restart the system and let it run without emitting light for 2 minutes. 4. Call Palomar service. 1. Ensure that the system has adequate space for air flow. Make sure that there are no air restrictions or blocked vents. 2. Let the system cool 3. Restart the system and let it run without emitting light for 2 minutes. 4. Call Palomar service. 1. Restart the system. 2. Call Palomar service. 1. If temperature is unstable or operating abnormally, replace power supply module. 1. Monitor 24Vdc on ThermoController. If too high, then replace power supply module. 1. Monitor input voltage from PFC. If greater than 410Vdc then, check PFC output. 2. Replace power supply module. 1. Restart the system. 2. Call Palomar service. 1. Replace power supply module 2. Call Palomar service. 1. 2. 1. 2. 1. 2. 3. 1. Replace power supply module. 2. Call Palomar service. 1. Restart the system. 2. Call Palomar service. Restart the system. Call Palomar service. Restart the system. Call Palomar service. Restart the system. Try another handpiece. Call Palomar service. 001- 66 Simmer Error. Internal LampController voltages exceed limits. Internal LampController voltages exceed limits. 001- 67 Lamp Current exceeds limits. Restart the system. Try another handpiece. Call Palomar service. Restart the system. Try different operating set points. 3. Try another handpiece. 4. Call Palomar service. 68 The system is not able to establish simmer. 1. Restart the system. 2. Try another handpiece. 3. Call Palomar service. 001- 001- 65 Chapter 3 P/N 5500-0000 rev X2 1. If temperature is unstable or operating abnormally, replace power supply module. 1. Restart the system. 2. Try another handpiece. 3. Call Palomar service. 1. 2. 3. 1. 2. 1. Replace power supply module. 1. Replace handpiece. 2. Replace power supply module. 1. Replace handpiece. 2. Replace power supply module. 1. Replace handpiece. 2. Replace power supply module. 1. Replace power supply module. 2. Replace handpiece. 1. Check High Voltage pins on connector. 2. Replace handpiece. 3-7 Technical Service Manual StarLux500 001- 69 001- 70 Internal LampController voltages exceed limits. Internal LampController voltages exceed limits. 001- 72 001- 73 Internal LampController temperature exceed o 60 C. Internal LampController temperature sensor is disconnected. Broken wire inside umbilical. High voltage pin is pushed out. Internal wiring disconnected. 001- 74 Internal wiring disconnected. 75 Operating parameters exceeds limits. 76 Operating parameters exceeds limits. 77 Operating parameters exceeds limits. 001- 001- 001- 001- 71 001- 78 001- 79 Operating parameters exceeds limits. Internal Communications error. 001- 80 81 System flashes at too high Rep. Rate. (Hz) Not used. 001- Chapter 3 P/N 5500-0000 rev X2 1. Restart the system. 2. Call Palomar service. 1. Restart the system. 2. Call Palomar service. 1. Ensure that the system has adequate space for air flow. Make sure that there are no air restrictions or blocked vents. 2. Let the system cool 3. Restart the system. Let system run without emitting light for 2 minutes down. 4. Call Palomar service. 1. Let the system cool down. 2. Restart the system. 3. Call Palomar service. 1. Replace the handpiece. 2. Call Palomar service. 1. 2. 1. 2. 3. 4. 1. 2. 3. 4. 1. 2. 3. 4. 1. 2. 3. Restart the system. Call Palomar service. Restart the system. Try different operating set points. Try another handpiece Call Palomar service. Restart the system. Try different operating set points. Try another hand piece. Call Palomar service. Restart the system. Try different operating set points. Try another handpiece. Call Palomar service. Restart the system. Try different operating set points. Try another handpiece. 1. Restart the system. 2. Call Palomar service. 1. Treat at slower speeds. Not used. 1. Check Voltage on main capacitors. If voltages are low, 2. Check Charger output. 3. Replace power supply module. 1. Check Voltage on main capacitors. If voltages are high, check Charger output. 2. Replace power supply module. 1. If temperature is unstable or operating abnormally, replace power supply module. 1. If temperature is unstable or operating abnormally. 2. Replace power supply module 1. Replace handpiece. 2. Check/replace high voltage pins on the connection assembly. 3. Replace connection assembly. 1. Verify Cooling Interlock cable is connected. 2. Replace power supply module. 1. Replace handpiece. 2. Install new compact flash card with software. 1. Replace handpiece. 2. Install new compact flash card with software. 1. Replace handpiece. 2. Install new compact flash card with software. 1. Replace handpiece 2. Install new compact flash card with software. 1. Replace power supply module. 2. Replace display module. 1. Replace compact flash card with new software. 2. Replace display module. 3. Replace power supply module. Not Used. 3-8 Technical Service Manual StarLux500 001- 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. Not used. 001- 97 Internal PFC communication error. 1. Restart the system. 2. Call Palomar service. 98 Internal ThermoController communication error. 1. Restart the system. 2. Call Palomar service. 001- 99 100 101 Internal LampController communication error. Not used. Not used. 1 Restart the system. 2 Call Palomar service. Not used. Not used. 001- 102 Internal PFC communication error. 1 2 Restart the system. Call Palomar service. 103 Internal ThermoController communication error. 1 2 Restart the system. Call Palomar service. 104 105 106 Internal LampController communication error. Not used. Not used. 1. Restart the system. 2. Call Palomar service. Not used. Not used. 001001001001001001001001001001001001001001- 001- 001001- 001- 001001001- Chapter 3 P/N 5500-0000 rev X2 Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. Not Used. 1. Check for loose connections to power supply module and display module. 2. Replace power supply module. 3. Replace display module. 1. Check for loose connections to power supply module and display module. 2. Replace power supply module. 3. Replace display module. 1. Check for loose connections to power supply module and display module. 2. Replace power supply module. 3 Replace display module. Not used. Not used. 1. Check for loose wiring in power supply module and display module. 2. Verify PFC/Charger has latest revision. 3. Replace power supply module. 1. Check for loose wiring in power supply module and display module. 2. Verify PFC/Charger has latest revision. 3. Replace power supply module. 1. Check for loose wiring in power supply module and display module. 2. Verify PFC/Charger has latest revision. 3. Replace power supply module. Not Used. Not Used. 3-9 Technical Service Manual StarLux500 1 001- 107 Internal System communication error. 2 3 1. 001- 108 Internal hand piece communication error. 2. 3. 1. 109 Internal hand piece communication error. 001- 110 Internal compact flash card access error. 2. 3. Ensure good hand piece connection. Restart the system. Call Palomar service. Ensure good handpiece connection. Restart the system. Call Palomar service. Ensure good hand piece connection. Restart the system. Try another handpiece Call Palomar service. Ensure that the compact flash card is inserted properly. Restart system. Call Palomar service. 001- 111 001- 112 Internal System communication error Internal System operating state error. 1. 2. 1. 2. Restart the system. Call Palomar service. Restart the system. Call Palomar service. 113 Incorrect Internal PFC software version. 1. Restart the system. 2. Call Palomar service. 114 Incorrect Internal ThermoController software version. 1. Restart the system. 2. Call Palomar service. 001- 115 116 Incorrect Internal LampController software version Not used. 1. Restart the system. 2. Call Palomar service. Not used. 001- 117 Not used. Not used. 001- 118 Internal System communication error 1. Restart the system. 2. Call Palomar service. 119 System Hardware/Software versions are incompatible. 1. Restart the system. 2. Call Palomar service. 001- 001- 001- 001- 001- Chapter 3 P/N 5500-0000 rev X2 2. 3. 4. 1. 1. Check for loose wiring in power supply module and display module. 2. Replace power supply module. 3. Replace display module. 1. Replace handpiece. 1. Replace handpiece. 1. Install new compact flash card with software. Don’t swap CF cards between systems. 2. Replace display module. 1. Check for loose wiring in power supply module and display module. 2. Replace power supply module. 3. Replace display module. 1. Replace display module. 1. If display module recently replaced verify revision of power supply module 2. Replace display module. 3. Replace power supply module. 1. If display module was recently replaced verify revision of power supply module. 2. Replace display module. 3. Replace power supply module. 1. If display module recently replaced verify revision of power supply module. 2. Rplace display module. 3. Replace power supply module. Not Used. Not Used. 1. Check wiring for loose connections in power supply module and display module. 2. Replace power supply module. 3. Replace display module. 1. If the software was just replaced verify revisions of modules. 2. If a module was just replaced verify revision of softwar.e 3. Replace software. 4. Replace display module. 3-10 Technical Service Manual StarLux500 001- 120 001- 121 001- 122 001- 129 Internal wiring became disconnected. 1. Restart the system. 2. Call Palomar service. 1. 2. 1. 2. 001- 125 001- 001- Try another handpiece. Call Palomar service. Restart the system. Call Palomar service. 128 123 124 001- 1. 2. 1. 2. Error occurred during flashing. The system delivered energy to the flashlamp that exceeded limits. The system delivered current to the flashlamp that exceeded limits. The system delivered peak power to the flashlamp that exceeded limits. The system delivered peak power to the flashlamp that exceeded limits. The system delivered current to the flashlamp that exceeded limits. The system delivered current to the flashlamp that exceeded limits. 001- 001- Handpiece Hardware/Software versions are incompatible. Error occurred during flashing. 126 127 1. Contact Palomar Service. 1. Restart system. 2. Try another hand piece. 3. Call Palomar service. 1. Contact Palomar Service. 1. 2. 3. 1. Restart system. Try another handpiece. Call Palomar service. Restart system. If problem persists, 2. Try another hand piece. 3. Call Palomar service. 1. Contact Palomar Service. 1. Contact Palomar Service. 1. Restart system. 2. Try another handpiece. 3. Call Palomar service 1. Contact Palomar Service. 1. Restart system. 2. Try another handpiece. 3. Call Palomar service. 1. Contact Palomar Service. 1. Restart system. 2. Try another handpiece. 3. Call Palomar service. 001- 130 001- 131 001- 132 Error occurred during flashing. 1. Restart the system. 2. Call Palomar service. 133 Internal Voltage loaded down. 1. Restart system. 2. Try another handpiece. 3. Call Palomar service. Chapter 3 P/N 5500-0000 rev X2 1. Contact Palomar Service. 1. Restart the system. 2. Call Palomar service. Internal wiring became disconnected. Error occurred during flashing. 001- 1. If the software was just replaced verify revisions of modules. 2. If a module was just replaced verify revision of software. 3. Replace software. 4. Replace display module. Restart the system. Call Palomar service. Restart the system. Call Palomar service. 1. Contact Palomar Service. 1. Check wiring for loose connections in power supply module and display module. 2. Replace power supply module. 3. Replace display module. 1. Check wiring for loose connections in power supply module and display module 2. Replace power supply module. 3. Replace display module. 1. Contact Palomar service. 1. Contact Palomar service. 1. Replace handpiece. 2. Check/replace connection plate. 3. Replace power supply module. 4. Replace display module. 3-11 Technical Service Manual StarLux500 001- 001001001- 001- 001- 001- 001- 001- 001- 134 Internal Voltage loaded down. 1. Restart system. 2. Try another handpiece. 3. Call Palomar service. 135 136 137 PFC temperature exceeds limits. Not used. Not used. 1. Let the system cool down. 2. Restart the system. 3. Call Palomar service. Not used. Not used. 138 139 140 141 142 143 1064 HP board temperature. 1. Cool down system for 15 minutes. 2. Restart system. 3. Try another handpiece. 4. Call Palomar service. 1. Replace handpiece. 2. Check/replace connection plate. 3. Replace power supply module. 4. Replace display module. 1. Monitor PFC temperature sensors. If temperature is unstable or operating abnormally, replace power supply module. Not used. Not used. 1. Using System Debug Software, check that the handpiece temperature is o o between 10 & 50 C and o o remains between 10 & 50 C for 20 minutes. 2. If temperature is wrong, the handpiece board needs to be replaced. 1. Verify that software CF card is the latest revision. 2. If system is over s/n 151873, the user may be trying to install a handpiece that is not allowed to be operated on the system. 3. Call Palomar service. 1. Install upgrades required for the operation with the new handpiece type. 2. Contact Palomar service. 1. Restart system. 2. Try another handpiece. 3. If problem re-occurs with IR handpiece, call Palomar service. 1. Restart system. 2. If problem, re-occurs with IR handpiece, replace handpiece. 1. Restart system. 2. Try another handpiece. 3. Call Palomar service. 1. Restart system. 2. Try another handpiece. 3. Contact Palomar service. 1. Verify the IR handpiece internal cooling fan is operating. 2. Restart system. 3. Call Palomar service. 1. Let system cool down for 5 minutes. 2. Restart system. 3. Verify the IR handpiece internal cooling fan is operating. 4. Return IR handpiece for repair. 1. Verify a 1540 tip is attached properly. 2. If problem re-occurs, return handpiece for repair. 1. Remove the handpiece tip and re-install. 2. Restart system. 3. Contact Palomar service. HP not supported. IR lamp controller did not receive a signal from the handpiece. IR handpiece data. IR handpiece over temperature. Erbium HP Tip Disconnected. Chapter 3 P/N 5500-0000 rev X2 3-12 Technical Service Manual StarLux500 001- 001- 144 145 Erbium HP Tip Not Supported. External 500 Chiller Unit fault. Chapter 3 P/N 5500-0000 rev X2 1. Install a supported 1540 handpiece tip. 2. Restart system. 3. Call Palomar Service. 1. Verify that a supported 1540 2. Restart system. 3. Return handpiece for repair. 1. Make sure the power cord and serial communication cable to the external chiller are connected. 2. Verify the power switch to the chiller is on. 3. Restart system. 4. Call Palomar service. 1. Verify all connections to and from the chiller. 2. Restart system. 3. Contact Palomar service. 3-13 Technical Service Manual StarLux500 3.1.1 Non-Error Code Faults Symptom Blank Display or Horizontal/Diagonal Lines across Display. Possible Cause Voltage Regulator overheating. Customer Correction 1. Turn off the power switch at the rear of the system. 2. Let system cool down for at least 5 minutes. 3. Restart system. 4. If problem goes away, turn off the power switch in the rear of the system when not in use. 5. Replace Display Module and software. Start.exe Software Error. The start program has failed causing an internal system fault 1. The system will rebuild the files in about 15 minutes. Do not turn the system off at during this time. 2. If problem persists, replace Display Module and software. Hourglass symbol above Thermometer. Release Footpedal. The handpiece temperature is exceeding parameters and the user is trying to operate the footpedal while the hourglass is still present on the screen. After releasing the footpedal, let the system cool for 10 minutes. Do not turn the system off. (symbol looks like a laptop) System is cooling down the handpiece. The cool down time during normal operation and room temperature can take up to 7 minutes. Cool down could be longer if room temperature o is above 75 F. LuxIR handpiece may take up to 20 minutes. Software Compact Flash Card error. 1. Verify the CF card is installed with the ridge of the card facing down. 2. If problem persists, replace Display Module and software. Replace Display Module and software. System is randomly Rebooting itself. Display failure. Chapter 3 P/N 5500-0000 rev X2 3-14 Technical Service Manual StarLux500 Footpedal not working or stops pulsing hp. while Starlux Air Switch or Footpedal problem. Seen only on 1064 and 1540 handpieces. Screen displays a Door opening and closing. Verify that footpedal is securely connected to back of system. Can also cut end of black hose inch with scissors and reconnect. If problem persists, check with Palomar Service before replacing footpedal or system air switch. Lamp warning message after calibration. OK to continue using handpiece. The problem may have been the Calibration tip was loose (on Lux1540). Note: Handpiece may begin failing next calibration and require factory adjustments. Remote Interlock is not connected or loose. This graphic will only appear while using a Lux1064 or Lux1540 laser handpiece. Internal Flash Memory Corrupted – Software Error. 1. Leave StarLux on until Auto -Recovery is complete. System may reboot itself. 2. If the system does not recover after several minutes, try restarting the system. 3. If screen does not return to normal, replace Display Module and software card. StarLux500 systems Chapter 3 P/N 5500-0000 rev X2 1. 2. 3. 4. External Cooling Module is not turned on. Communication Cable not connected. Check Cable Connection, restart system. Replace the External Cooling Module. 3-15 Technical Service Manual StarLux500 Chapter 4 Service Procedures 4.1 Verification of Energy 4.1.1 Test Equipment Requirements 4.1.2 Verify Energy 4.1.3 Verification of Pulsewidth 4-2 4-2 4-2 4-2 4.2 Debug Software Operation 4.2.1 Main User Screen 4.2.2 System Controller Screen 4.2.2.1 Confirming System Cooling 4.2.2.2 Verifying Pump Operation 4.2.2.3 Verify Charge and Simmer 4-3 4-3 4-4 4-4 4-5 4-6 4.3 System Error Log Utility 4-7 4.4 Exporting the Error Log 4.4.1. Error Log Export 4.4.2 Printing Error Log 4-8 4-8 4-8 4.5 Clearing the Error Log 4-9 4.6 System Configuration Utility 4.6.1 Replacing the Display Module 4-9 4-9 4.7 Re-Entering System Information 4.7.1 Editing System Type/Serial Number 4.7.2 Editing Manufacture/Warranty Date 4.7.3 Clearing System Shot Count 4.7.4 Exiting System Utility Screen 4-10 4-10 4-10 4-11 4-11 4.8 HP Reader Utility 4.8.1 HP Reader Utility Screen 4.8.2 HP Serial Number and Shot Count 4.8.3 Exiting HP System Utility Screen 4-11 4-11 4-12 4-12 Chapter 4 P/N 5500-0000 rev X2 4-1 Technical Service Manual StarLux500 4.1 Verification of Energy Energy should be verified based on local requirements or customer’s request. All calibration data is stored in the handpiece. Therefore each Starlux500 system has no calibration requirement. 4.1.1 Test Equipment Required • Ophir L40(150)A-SH-V2 Power Meter Head. • Ophir Nova Display. • ThorLabs DET10A Optical Sensor. • Oscilloscope (ie. Fluke123 Scopemeter) • Stylus Pen (for display touch screen) • Starlux500 LuxR Test Handpiece 4.1.2 Verify Energy 1. Attach handpiece with a large surface such as a 500 LuxR. 2. Adjust the system settings following the chart below. 3. Pulse the system and average 5 shots per energy range. Please Note: The Energy cannot be adjusted on a Starlux500. Pulsewidth (ms) Fluence (J/cm2) Expected Energy (J) L40(150)A-SH-V2 5 20 60 100 2 5 6 11 Min 7.3 18.3 21.9 40.2 Max 8.9 22.3 26.8 49.1 4.1.3 Verification of Pulsewidth 1. Select the lowest fluence available for the pulse width selected. 2. Pulse the system with the handpiece toward the photodetector and record the pulse width. Compare to chart below. Please Note: The Pulsewidth cannot be adjusted on a Starlux500. SETTING 10 ms 20 ms 100 ms Chapter 4 P/N 5500-0000 rev X2 RANGE 9 – 11 ms 18 – 22 ms 90 – 110 ms 4-2 Technical Service Manual StarLux500 4.2 Debug Software Operation The system is designed to be operated with software that access specific components to be turned on individually. This will allow for troubleshooting the system in non-operator mode. Remove the customer compact flash card and insert Debug Software into customer system. Please Note: The Keyswitch and E-Stop are disabled during debug operation. Figure 1 – Main User Screen 4.2.1 Main User Screen Figure 1 shows an example of the Main User screen for System Debug software. Each utility shown can be accessed through their corresponding command buttons from the Main User screen. Using the optional Stylus Pen will make it easier to touch the screen buttons. To enter service mode, attach a Starlux500 LuxR test handpiece, select IPL 500 HP and press System Controller Utility. Chapter 4 P/N 5500-0000 rev X2 4-3 Technical Service Manual StarLux500 Step 4 Step 6 & 7 Step 5 Step 2 Step 3 System Controller screen 4.2.2 System Controller Screen 4.2.2.1 Confirming System Cool Down Operation: 1. Check the Reservoir for coolant to insure it is at full level. 2. Select PFC ON then DC_DC On to turn power supplies on. 3. Select Pumps On to switch Lamp and Cooltip pumps on. 4. Set Fan Speed % to 80 using the arrow buttons on left and then select Fan Set to run fans. 5. Select Polling to ON. This will display the active Lamp and HP temperature measurements as detected by the Starlux internal thermistors. 6. Set TEC Temp to 20o C using arrow buttons on left and let the system run for 1 - 2 minutes. 7. Set TEC Temp to 2o C and measure the time (in minutes), it takes to reach 5o C. This time interval should be less than 5 minutes. 8. When testing is completed, select EXIT to end the Debug program. Chapter 4 P/N 5500-0000 rev X2 4-4 Technical Service Manual StarLux500 Step 1 Step 2 System Controller Screen 4.2.2.2 Verifying Pump Operation: 1. Select PFC On then DC_DC On to turn power supplies on. 2. Select Pumps On to switch Lamp and Cooltip pumps on. 3. When testing is completed, select EXIT to end the Debug program. Chapter 4 P/N 5500-0000 rev X2 4-5 Technical Service Manual StarLux500 Step 3 Step 2 Step 1 Step 4 Step 5 System Controller Screen 4.2.2.3 Verifying Charge and Handpiece Simmer: 1. Select PFC On then DC_DC On to turn on the power supplies. 2. Select Fan Set to turn fans on full speed. (100%). 3. Select Pumps On to turn Lamp and Cooltip pumps on. 4. Select Charge On to enable high voltage charging. 5. Select Lamp On to initiate handpiece simmer. 6. When testing is completed, select EXIT to end the Debug program. Chapter 4 P/N 5500-0000 rev X2 4-6 Technical Service Manual StarLux500 4.3 System Error Log Utility The System Error Log Utility allows the user to read error log data from the system. Step 1 Main User Screen 1. Press the System Error Log Utility button from the Main User Screen. Step 2 Error Log Screen 2. Then select Display Error Log to display the error log data on the screen. Chapter 4 P/N 5500-0000 rev X2 4-7 Technical Service Manual StarLux500 4.4 Exporting the Error Log Debug Software version 1.1 and higher features an Export Error Log option which allows the user to copy the system’s error log data to the Debug card. Once the error log is downloaded, power down the Starlux500 before removing Debug. You can then install the Debug card to a CF card reader that is connected to a PC, and create a computer printout of the data. 4.4.1 Error Log Export Select the Export Error Log command button to have the error log data transferred to the Debug card installed. A message of “ERROR LOG SUCCESSFULLY EXPORTED!” is printed on the screen. Error Log Screen 4.4.2 Printing the Error Log A folder labeled with the system serial number is created on the root directory of the Debug compact flash card. (fig. 1) Within that folder is a text file labeled with the system serial number that contains the error log data as seen on the screen.(fig. 2) Open this file from your PC and print a hardcopy of the data. Figure 1. Error Log Folder Chapter 4 P/N 5500-0000 rev X2 Figure 2. Error Log Data File 4-8 Technical Service Manual StarLux500 4.5 Clearing the Error Log Select the ‘Clear Error Log / Create New Error Log’ command button to erase the error log data off the system. A message of “ERROR LOG CLEARED - NEW ERROR LOG CREATED.” is printed on the screen. Select the Close command button to exit the System Error Log Utility screen. 4.6 System Configuration Utility 4.6.1 Replacing the Display Module When replacing the Display on a Starlux500, the system information must be configured on the Compact Flash Software card from the Palomar factory. Reinstalling the original software card will not work on the new display and you will get a 001-110 error. Therefore you MUST provide the serial number of the Starlux500 before ordering and receiving a new software card with the new Display. The system information along with a Software Key will then be loaded and shipped with the new Display prior to your installation on the system. This Software Key (Password) can also be emailed. If you have any questions regarding this procedure, please contact or email Palomar Customer Service. Main User Screen Chapter 4 P/N 5500-0000 rev X2 4-9 Technical Service Manual StarLux500 4.7 Re-Entering System Information After selecting System Configuration Utility from the Main User Screen, the System Type, Serial Number, Manufacture Date, Warranty Date, and Shot Count will appear in their respective fields. If no system data is present, the System Type and Serial Number will have values of “Empty”. System Configuration Utility Screen 4.7.1 Editing System Type and Serial Number To change the System Type and System Serial Number values on a system, using the optional Stylus Pen, tap the text field you wish to edit. A little keyboard screen will pop up and allow the user to edit the text field selected. When all changes are complete, select the Save Configuration command button. Once the data has been saved successfully, the text field information you entered will be stored. Please Note: After installing a new Display, you may see a default value of “Starlux” and “P-0002” printed in the System Type and System Serial Number text fields respectively. The Manufacture Date and Warranty Date may default to the current date. This is normal until you enter the system information. 4.7.2 Editing Manufacture and Warranty Dates To change the Manufacture and Warranty Date values on a system, using the optional Stylus Pen, tap the text field you wish to edit. A calendar menu will pop up and allow the user to select a month, day and year. When all changes are complete, tap the screen right side edge to accept your entry or select the Save Configuration command button if finished. Once the data has been saved successfully, the text field information you entered will be stored. Chapter 4 P/N 5500-0000 rev X2 4-10 Technical Service Manual StarLux500 4.7.3 Clearing the System Shot Count To clear the system shot count, select the Zero Shot Count command button. A value of “0” will be printed in the Shot Count field. To save, select the Save Configuration command button. Once the data has been saved successfully, the text field information you entered will be stored. Please Note: System shot count data cannot be changed or added. 4.7.4 Exiting the System Configuration Utility Screen Select the Close command button to save all your changes and exit the System Configuration Utility. 4.8 HP Reader Utility The HP Reader Utility allows the user to read data stored on the HP Board. Main User Screen 4.8.1 HP Reader Utility Screen With a handpiece attached, select the HP Reader Utility command from the Main User Screen. If there is no HP attached to the system, these fields will have a value of “NO HANDPIECE DETECTED”. If there is no data on the HP connected, these fields will have a value of “NO EEPROM DATA”. Chapter 4 P/N 5500-0000 rev X2 4-11 Technical Service Manual StarLux500 HP Reader Utility Screen 4.8.2 HP Serial Number and Shot Count After selecting HP Reader Utility, the HP Serial Number and HP Shot Count fields are updated with the data from the HP board. The Display Cal Log, Display Error Log and Save Cal Log command buttons are not applicable to the user and will display the message “THIS OPTION IS NOT APPLICABLE FOR THIS HP TYPE”. Please Note: HP Serial Number and HP Shot Count data cannot be changed. 4.8.3 Exiting HP Reader Utility Select the Close command button to exit the HP Reader Utility screen. Chapter 4 P/N 5500-0000 rev X2 4-12 Technical Service Manual StarLux500 Chapter 5 Disassembly 5.1 Removing the Display 5-2 5.2 Removing the Power Supply 5-3 5.3 Removing the Capacitor Bank 5-5 5.4 External Cooling Unit 5.4.1 Starlux Thermisters 5.4.2 Cooling Unit Models 5.4.3 Cooling Unit Chassis 5.4.4 Cooling Unit Components 5.4.5 Cooling Unit Power Supply 5-5 5-5 5-6 5-7 5-7 5-8 5.5 Starlux500 System Carts 5-8 Chapter 5 P/N 5500-0000 rev X2 5-1 Technical Service Manual StarLux500 5.1 REMOVING THE DISPLAY MODULE 1. Turn system power off and disconnect power cord, cooling unit communication cable and hose assembly from Starlux500. 2. Remove all four (10-32) screws that secure the Bezel Assy to the chassis. (see fig. 1 & 2) Figure 1. (Right Side of Bezel) Figure 2. (Left Side of Bezel) 3. Carefully lift the Bezel Cover Assy from the chassis (Figure 3). Disconnect wires to the Keyswitch, Emergency Stop Button (Figure 4) and dock sensor magnets. (Figure 5) Then gently lift the Bezel Cover off. Figure 3. (Bezel Assy lifted) Figure 4. (Keyswitch wiring) Figure 5. (Dock Sensor wires) 4. Remove the 2 screws that hold the Rear Vent Assy (fig. 6) then remove all chassis screws (fig. 7). Figure 6. (Rear Vent Assy removal) Chapter 5 P/N 5500-0000 rev X2 Figure 7. (Top Lid mounting screws) 5-2 Technical Service Manual StarLux500 5. Lift the entire front panel forward to access the interior of the system (figure 8) and disconnect all cables from the rear of the display. (figure 9) Open Ferrite Block here if installed. Figure 8. (Top Lid opened) Figure 9. (Display rear panel connections) 6. Remove 3 phillips screws on the left securing the display module, then remove the 2 phillips screws on the right and lift the display module out. (Figure 10 & 11) Figure 10. (Display mounting screws) Figure 11. (Display removed) 5.2 REMOVING THE POWER SUPPLY 1. Remove the Bezel Cover Assy as instructed in steps 1 thru 5 in Section 5.1. 2. Disconnect the High Voltage cables from the Lamp Controller Board. (Fig. 13) Put a jumper across the +400V and -400V wires. (Fig. 14 ) CAUTION: Failure to short these wires prior to removing cables, will cause the High Voltage capacitors to “charge in air”. This will cause a potentially dangerous voltage on the capacitors! 3. Disconnect the two pin connector on the Charger Board that connects to the capacitors. 4. Remove all ribbon cables connected. 5. Disconnect the 4 pin cable from the Voltage Regulator Board located on the side of the power supply. Chapter 5 P/N 5500-0000 rev X2 5-3 Technical Service Manual StarLux500 HV Cables Figure 12. (System Overview) Figure 13. (Power Supply connections Jumper Wires Figure 14. (Capacitor Jumpers) Figure 15. (Power Supply mounting screws) Figure 16. (removing Power Supply) Figure 17. (Power Supply removed) 6. Remove the 3 rear mounting screws of the Power Supply indicated by the black circle. (Fig. 15) 7. Pull the Power Supply from the rear and out of the Starlux rear chassis (Fig.16 and 17) 8. To reinstall Power Supply, reverse steps 1 thru 8. Chapter 5 P/N 5500-0000 rev X2 5-4 Technical Service Manual StarLux500 5.3 REMOVING THE CAPACITOR BANK 1. Remove Bezel Cover Assy as instructed in steps 1 thru 5 in Section 5.1. 2. Remove the Power Supply as instructed in steps 1 thru 8 in Section 5.2. Disconnect the High Voltage cables from the Lamp Controller Board. (Fig. 13) Put a jumper across the +400V and -400V wires. (Fig. 14 ) CAUTION: Failure to short these wires prior to removing cables, will cause the High Voltage capacitors to “charge in air”. This will cause a potentially dangerous voltage on the capacitors! 3. Loosen and remove the two Phillips screws holding the front of the capacitor bracket sheet metal to the chassis. (Figure 18) 4. Check the opposite end of the capacitor bracket. If there are screws holding the bracket down, loosen them. If the sheetmetal bracket has tabs that slide under the main chassis, there are no screws to loosen. 5. Pull up on the front edge of the capacitor bracket and gently pull the assembly forward and up to remove it.(Figure 19) Take care not to damage the speaker assembly mounted under the capacitors when removing the assembly. Figure 18. (Capacitor Bank mounting screws) Figure 19. (Capacitor Bank removed) 5.4 EXTERNAL COOLING UNIT The External Cooling Module provides cooling for Starlux500 base unit for all it’s handpieces. It uses a 4 line hose assy to connected to the Starlux, a serial communication cable and an AC power cord. 5.4.1 The Starlux 500 contains a primary heat exchanger and two Thermisters located at the rear of the base unit chassis below the Reservoir. (see photo) Lamp HP Starlux 500 Chassis Thermisters Chapter 5 P/N 5500-0000 rev X2 5-5 Technical Service Manual StarLux500 5.4.2 Starlux500 Cooling Unit Models Connector to Cart Bezel Cable Cooling Unit 1520-5008 Control Panel for 1520-5008 Cooling Unit 1520-5034 (released on 7/07) Control Panel for 1520-5034 Cooling Unit Hose Assembly Cooling Unit Hose Connection Panel Chapter 5 P/N 5500-0000 rev X2 5-6 Technical Service Manual StarLux500 5.4.3 Cooling Unit Chassis (opened) 5.4.4 Cooling Unit Internal Components HP Pressure Switch Secondary Heat Exchanger Lamp Pressure Switch HP Pump Junction Board Aquatec Pump Thermostat location for each TEC. Chapter 5 P/N 5500-0000 rev X2 TEC’s 5-7 Technical Service Manual StarLux500 5.4.5 Cooling Unit Power Supply Chassis Adapter Board Control Board +24Vdc Power Supply 5.5 Starlux500 System Carts Original 500 Cart 1520-5009 Chapter 5 P/N 5500-0000 rev X2 500 Cart 1520-5020 (released 7/07) 5-8