Data Sheet EI Series Bimetal Thermometers SPECIFICATIONS Ashcroft® Series: EI Dial Sizes: 2,˝ 3,˝ 5˝ Stem Length: 21⁄2˝-60˝ (1) Case & Stem: 304 SS, hermetically sealed Stem Dia: .250˝ Window: Heavy-duty glass, plastic or shatterproof glass optional Dial: Maxivision®, black figures on white background Pointer: Black Connection: Plain, pointed plain, 1⁄4 NPT, 1⁄2 NPT, 1 ⁄2 NPT union and 1⁄2 NPT adjustable union Connection Location: Everyangle™, Lower, Rear Ranges: –80/120°F-200/1000°F –50/50°C-100/500°C Options: Code Description C4 Individual calibration cert. XCS Dual scale(2) XDM Dial marking XNH SS tag XNN Paper tag XPD Plastic window XSG Shatter proof glass X3B 3⁄8˝ stem dia. with 1⁄2 NPT X02 1⁄4 NPT when 1⁄2 NPT is standard(3) XS1 Silicone free XYW 316 SS Housing(4) and stem FEATURES • 1% full span accuracy ASME B40.200 (ASME B40.3 Grade A) • Maximum ambient temp. is 200°F (94°C) • Hermetically sealed case to prevent entry of moisture, interior corrosion and coil freeze-up. • External adjustment permits zero reset from outside the case. • Maxivison® dial allows readability from any angle without parallax error. • Silicone coil dampening (up to 400°F) provides vibration dampening and improves response time. • All-welded SS construction • Heavy-Duty glass lens • Protection NEMA 4X/IP66 • Five-year limited warranty NEW L A OPTIONUSING (1) 25˝-60˝ lengths available, consult factory (2) Dual scale available with 3˝ and 5˝ case only (3) Only available on rear connection (4) Only available on 5˝ Everyangle™ HO 316SS EM & ST 5˝ BLE ON AVAILA NGLE EVERYA Thermowells must be used on all pressure or velocity applications, to protect the stem of thermometer from corrosion and physical damage, and to facilitate removal of the thermometer without disturbing the process. Maximum ambient temperature is 200°F (95°C). HOW TO ORDER 30 EI 60 R 040 0/250°F XNH Dial Size: 3" Code 30 Case Style: EI Stem Connection: 1/2 NPT Code 60 Stem Location: Rear Code R Stem Length: 4” Code 040 Range: Code 0/250°F Options: Stainless steel tag BULLETIN BM-EI All specifications are subject to change without notice. All sales subject to standard terms and conditions. © 2017 Ashcroft Inc. 07/17 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA Tel: 203-378-8281 • Fax: 203-385-0408 email: info@ashcroft.com • www.ashcroft.com Data Sheet EI Series Bimetal Thermometers EI Series dual scale ranges EI Series selection table Case Size Dial Code 2˝ Stem Style Code 20 3˝ 30 Conn. Code Location Plain Pointed 1 /4 NPT 1 /2 NPT Union 1 /2 NPT 40 50 60 Rear /2 NPT 60 1 /2 NPT Union 1 /2 NPT 1 /2 NPT Adjust. Union 1 /2 NPT EI 5˝ 1 50 /2 NPT 1 42 Stem Lengths Available “S” Code Length Code (inches) °F* Fahrenheit –80/120 –20/120†† –40/120 30/130†† 0/200 20/120 0/250 50/300 50/400 50/550 200/700† 100/800† 200/1000**† R Everyangle E Rear Lower R L 60 42 60 Everyangle 21/2 4 6 9 12 15 18 24 Dual scales are available in 3" and 5" dial sizes in the following ranges: Temperature Range 025 040 060 090 120 150 180 240 E 70 °/Div. Fig. Inter. °C Celsius 2 20 –50/50 –10/110 –20/120 0/50†† 0/100 10/150 0/200 0/300†† 0/400† 50/450**† 100/500**† 1 10 20 2 50 5 50 10 100 Fig. °/Div. Inter. 1 2 2 1 1 2 2 5 5 5 5 Inner 10 10 20 5 10 20 20 50 50 50 50 60 Rear Lower 60 R L Outer –80/120°F and –60/50°C –40/120°F and –40/50°C –40/160°F and –40/70°C –20/120°F and –30/50°C†† 0/200°F and –20/90°C 0/250°F and –20/120°C 30/130°F and 0/55°C†† 50/300°F and 10/150°C 50/400°F and 10/200°C 50/550°F and 10/290°C 100/800°F and 50/400°C† 200/700°F and 100/370°C† 200/1000°F and 100/550°C**† CONTACT FACTORY FOR ADDITIONAL STEM LENGTHS AND RANGES NOT LISTED Overtemperature limits EI Dimensions Connection Case Dial Location Series Rear EI 2˝ (Plain) Rear (Plain, EI 2˝ pointed stem) Rear EI 2˝ (Threaded) EI EI EI EI EI EI 3˝ Rear 3˝ Lower 3˝ Everyangle 5˝ Rear 5˝ Lower 5˝ Everyangle Weight in ounces(3) A 2 /32 (53) 23/32 (53) 23/32 (53) 35/32 (80) 35/32 (80) 35/32 (80) 51/32 (128) 51/32 (128) 51/32 (128) 3 B /8 (10) 3 /8 (10) 3 /8 (10) 19 /32 (15) 127/32 (47) 19 /32 (15) 23 /32 (18) 115/16 (49) 23 /32 (18) 3 C /8 (8) 3 /8 (8) 3 /8 (8) 3 /8 (8) Hex S – 21⁄2˝ /16 41/2 /16 41/2 /16 41/2 7 /8 7 /2 7 /8 11 1 /2 7 /8 10 –(2) 1 /2 7 /8 16 D E G H S NPT – – – – –(2) – 11 – – – – –(2) – 11 – – – 1 /4 –(2) 41/2 11 – – – 1 /2 –(2) 1 /4 2 /8 (67) /4 (6) – 1 /2 –(2) 1 – 121/32 (42) 31/32 (89) –(2) – – Top of Range °F Maximum Over Temperature up to 250 100% of span 251/550 50% of span 551/1000 800°F ** 3 5 – 1 – – /8 (8) – – – 3 /8 (92) /4 (6) – – –(2) 1 /2 7 /8 26 – 17/8 (48) 321/32 (93) –(2) 1 /2 7 /8 25 *Dual scale ranges available for all standard °F ranges (3˝ and 5˝ F case only) **Satisfactory for continuous service up to 800°F or 425°C. Can be used for intermittent service from 800 to 1000°F, or 425 to 500°C. Use Ashcroft Duratemp® thermometers for ranges above and below those listed above. †Minimum stem length for these ranges is 4˝. ††Minimum stem length for lower connection and Everyangle is 4˝. 3 5 – – 1 NOTES: 1. Figures in parenthesis ( ) are in millimeters. All other dimensions are in inches. 2. Standard “S” dimensions are 21⁄2, 4, 6, 9, 12, 15, 18 and 24 inches. Standard stem diameter is 1⁄4 inch. 3. Add 1 oz. for every 2 inches of stem length. DIMENSIONS G G B B S B A A A H A H D 1.00" S B BULLETIN BM-EI S S C 30EI60E 50EI60E 50EL60E 30EI42E 50EI42E 50EL42E All specifications are subject to change without notice. All sales subject to standard terms and conditions. © 2017 Ashcroft Inc. 07/17 E Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA Tel: 203-378-8281 • Fax: 203-385-0408 email: info@ashcroft.com • www.ashcroft.com Screwed Thermowells Model SM, ST, RT and HT Features Solid bore construction Materials according NACE MR 01-75/ISO 15156 Selection of materials Standard and customer specifications ISO 228-1 and ANSI/ASME B1.20.1 connections Ratings from 620 bar at 20 °C to 175 bar at 650 °C Applications Chemical and petrochemical industry Machine and apparatus construction Food and beverage industry Pulp and paper industry Technical specification Solid bore design Construction shank Lagging Standard U-length Bore diameter SM, ST in inch in mm in inch in mm Pressure rating at process connection in inch °F °C 70 21 200 93 400 204 600 316 800 427 1000 538 1200 649 Process connection Instrument connection Material Certifications & tests Accessories, options Rev. G 06/2016 RT LRT HT LHT Straight Stepped Tapered No Yes No Yes No Yes 1 5/8 2½ 4½ 7½ 10 ½ 13 ½ 16 ½ 22 ½ 41 63 114 191 267 343 419 572 Optional customer specification 0,260 0,385 1) 6,6 9,8 1) 7,0 10,5 1) 12,5 1) not for types RT and LRT Others on request, see coding table Depending on process connection (½", ¾" or 1") and bore diameter (0,260"/6,6 mm) for standard thermowell material stainless steel 316L (1.4404) (These data don’t apply to other materials) ¾, 1 ½ ¾, 1 PSI BAR PSI BAR PSI BAR PSI BAR 6500 448 6000 413 6000 413 9000 620 5500 379 5200 358 5200 358 7700 531 5000 345 4500 310 4500 310 6700 462 4000 275 4000 275 4000 275 5800 400 3800 262 3500 241 3500 241 5300 365 3500 241 3300 227 3300 227 5000 345 2500 172 2500 172 2500 172 3700 255 Bore 0,385 (9,8 mm) 40 % less G ½ A male, G ¾ A male, G 1 A male according ISO 228-1 ½ NPT male, ¾ NPT male, 1 NPT male, 1 ¼ NPT male according ANSI/ASME B1.20.1 Others on request ½ NPSM female standard, optional ½ NPT female or G ½ female Stainless steel 304 (1.4301), stainless steel 304L (1.4306), Stainless steel 316 (1.4401), stainless steel 316L (1.4404), stainless steel 316Ti (1.4571) Hastelloy B, Hastelloy C, Monel, Duplex, Nickel Other materials or coatings on request Material certificate 3.1 to EN 10 204, hydrostatic test, dye penetration test, X-ray, hardness test, NACE/ISO 15156 certificate Thermometers, Pt-100, thermocouples All specifications are subject to change without notice. G2.TWSCR/E LSM, LST General dimensions in inch (mm) S P Type: (G, NPT) SM, ST, RT, HT ¾ SM, - 2-½ 63 1-5/8 41 1 SM, - ½ - ST, -, HT 4 102 2-½ 63 ¾ - ST, -, HT 1 - ST, -, HT ½ *, RT, HT 6 152 4-½ 114 ¾ - ST, RT, HT 1 - ST, RT, HT ½ *, RT, HT 9 229 7-½ 191 ¾ - ST, RT, HT 1 - ST, RT, HT ½ *, RT, HT 12 305 10-½ 267 ¾ - ST, RT, HT 1 - ST, RT, HT ½ *, RT, HT 15 381 13-½ 343 ¾ - ST, RT, HT 1 - ST, RT, HT ½ *, RT, HT 18 457 16-½ 419 ¾ - ST, RT, HT 1 - ST, RT, HT ½ *, RT, HT 24 610 22-½ 572 ¾ - ST, RT, HT 1 - ST, RT, HT For LRT, LST, LHT: S = 1 ¾‘‘ + T + U - ¼‘‘ * = on request For LSM: S = 1 1/16‘‘ + T + U - ¼‘‘ inch Lagging „T“ T inch mm 1-½ 2 3 U inch = 1-5/8 = 2-½ 38 50 76 4-½ For all types P (G, NPT) d B / B* ½ inch 0,26 mm 6,6 5 inch /8 / 5/8 ¾ ¾ 0,26 0,385 6,6 9,8 ¾ - 1 0,26 1 0,385 - = not available 6,6 9,8 7 - H (G / NPT) D / D* mm 16 / 16 ½ / 7/8 / 7/8 19 / / 22 22 ½ / 49 /64 / /8 / 1-1/16 / 1-1/16 22 / / 27 27 ½ / 49 /64 / / inch ½ mm 12,7 / 12,7 1-1/8 1-1/8 30 / mm 30 5 12,7 / 16 19,5 / 19,5 1-5/16 / 1-1/8 1-5/16 / 1-1/8 32 / 32 / 30 30 12,7 / 16 19,5 / 19,5 1-5/8 1-5/8 41 / 41 / 36 36 /8 49 /64 5 /8 49 /64 inch / / 1-3/8 / 1-3/8 U mm inch mm Order information Process U-length connection All connecInch dimension tions are male (0162) 1 5/8" (50) ½ NPT 1) (only type SM and LSM) (51) G ½ A 1) 1) (0250) 2 ½ " Instrument connection All connections are female Lagging Shank type Bore Material Options (=) No lagging (SM) Limit space straight Inch dimension (C) (NF) Tagging stamped (2) (260) 0,260" (CL) 304L (1.4306) Cap & chain stainless steel (BP) Buffed and polished (CD5) NACE certificate (L) With lagging (=) ½ NPSM (2) ½ NPT (0450) 4 ½" 1) (75) ¾ NPT (51) G ½ (0750) 7 ½" 1) (76) G ¾ A (10) 1 NPT (1050) 10 ½ " (standard 1-½,” 2” or 3” (38, 50 or 75 mm), see above table. Other lengths to be specified in inch or mm) (only 1) U = 1 5/8") (ST) Straight (HT) Tapered (1350) 13 ½" 1) (12) 1-¼ NPT (1650) 16 ½" 1) 1) not for SM and bore 0,385 inch/9,8 mm others on request MM dimension (... mm) length in mm others on request others on request Standard available 6,6 mm 7,0 mm 8,5 mm 1) 9,8 mm 1) 10,5 mm 1) 12,5 mm 1) not for type SM, LSM 1) others on request only bore 0,260”, 6,6 or 7 mm (S) (9,8 mm) MM dimension (... mm) (2250) 22 ½" 1) 1) (385) 0,385" 1) (RT) Stepped 1) 1) (11) G 1 A (6,6 mm) 1) 304 (1.4301) 316 (1.4401) (SL) 316L (1.4404) (W2) (STI) 316Ti (1.4571) Dye penetration test (MQ) PMI test (M) Monel (YR…) X-ray photo (H) Hastelloy C Material certificates acc. DIN EN 10 204 (G) Hastelloy B (CD2) (N) Nickel (C3) (J) Duplex Hydrostatic internal pressure test not for type RT and LRT others on request others on request Lagging Shank type Bore diameter L HT 2.2 3.1 (W9) 100 bar 1 min. (W4) To customer specification others on request Order example Process Connection 75 Type U-length W Instrument connection 0250 = Ashcroft Instruments GmbH Germany Max-Planck-Str. 1, D-52499 Baesweiler P.O. Box 11 20, D-52490 Baesweiler Tel.: +49 (0) 2401 808-0 France „206“ ZA du Mandinet, 1/3 Rue des Campanules, F-77185 Lognes Tel.: +33 (0) 1 60 37 25 30 260 Website: www.ashcroft.eu Ashcroft Instruments Ltd. Unit 17 & 18 William James House Cowley Road, Cambridge CB4 0WX Tel.: +44 (0) 12 23 39 55 00 Material Options SL X NF e-Mail: sales@ashcroft.com Ashcroft ISTANBUL Gayrettepe Mah. Yildiz Posta Cad. Yildiz Residence No:24 K:1 D:4 34349 Besiktas –Istanbul Tel.: +90 (0)212 3270847