Atlas 100-30 Technical Manual 1-2 FLOOR, BUILDING 12, NO.28 YUHUA ROAD, AREA B, BEIJING TIANZHU AIRPORT INDUSTRIAL ZONE, BEIJING, 101318, PRC +86 10 6431 3433 FAX+86 10 6431 3185 Doc. 8660 Technical Manual Index Section 0 Doc. 9535 Page 1 of 65 Technical Manual INDEX General index (doc. 9535) Valid from 10th December 2008 Section 0 - Index Page Section 1 - General description 1 2 Introduction 8 1.1 Applications and used 8 1.2 Composition 8 1.2.1 Radiographic unit 8 1.2.2 Control panel 9 Technical Data 11 2.1 Classification of the device 11 2.2 Technical characteristics 11 2.2.1 Dimensions and weights 13 2.2.2 kV – mA Relationship 14 2.2.3 RX exposure time 15 Section 2 - Installation 1 2 3 Installation 18 1.1 Introduction 18 1.2 Electrical connections 19 Configuration 19 2.1 Changing the language for the unit console 19 2.2 Programming the anatomical techniques 20 Acceptance test 21 3.1 Checking cables and connectors 21 3.2 Electrical checks 21 Section 0 Page 2 of 65 Doc. 9535 Technical Manual INDEX Page Section 3 – Safety and Maintenance 1 2 Safety 26 1.1 Introduction 26 1.2 Responsibility declaration 27 1.3 Compliance and reference address 27 1.4 System safety 27 1.4.1 Mechanical safety 27 1.4.2 Electrical safety 27 1.5 Protection against ionising radiation 28 1.6 Residual risks 29 1.7 Signals 30 1.7.1 Symbols used 30 1.7.2 Labelling 30 Maintenance 31 2.1 Routine maintenance 31 2.1.1 General racommendations 31 2.1.2 Frequent checks and inspections 32 2.2 General checks and inspections 32 2.3 Cleaning and disinfection 32 2.4 Special maintenance 33 2.4.1 Troubleshooting 33 2.4.2 Replacement of components 34 2.4.3 List potentiometers 35 2.4.4 List leds 36 2.4.5 List fuses 37 2.5 Spare parts 38 2.6 Disposal of device 38 Section 0 Doc. 9535 Page 3 of 65 Technical Manual INDEX Page Section 4 – Use of the unit 1 Transport 40 2 Switching on 40 3 Emission of X-rays 43 3.1 Emission of X-rays using the manual technique 44 3.2 X-ray emission using the anatomical technique 46 Operations after use 47 4 Section 5 – X-ray setting 1 Introduction 50 1.1 Dip switch 51 2 Rotating anode stator power supply adjustment 51 3 Set-up procedure 52 3.1 Service menu 53 kV adjustment 54 4.1 Adjustment of the “set-up” voltage 54 4.2 Adjustment of maximum kV setting safety 54 4.3 Checking maximum kV setting safety 55 Adjustment of the filament current 56 5.1 Adjustment of filament current 57 6 mAs adjustment 59 7 Adjustment of the battery voltage 60 8 Checking the radiografic parameter 61 8.1 Checking kV and mAs during radiography 61 4 5 Section 0 Page 4 of 65 Doc. 9535 Technical Manual INDEX Section 6 – Drawings Annex 1 – Responsibility Declaration Annex 2 – Labelling Annex 3 – Test Sheet Manuals of unit components Declaration of conformity and test reports Section 0 Doc. 9535 Page 5 of 65 Technical Manual INDEX Section 0 Page 6 of 65 Doc. 9535 Technical Manual General description Section 1 Doc. 9535 Page 7 of 65 Technical Manual GENERAL DESCRIPTION 1 INTRODUCTION 1.1 APPLICATIONS AND USED The equipment is part of a RADIOGRAPHIC UNIT for radiography on X-Ray film, that may be used in different places and situations: operating theatre, orthopedics, intensive care, emergency room. 1.2 COMPOSITION 1.2.1 RADIOGRAPHIC UNIT 1 2 3 1 Control electronics assembly 2 Control panel 3 X-ray tube head Section 1 Page 8 of 65 Doc. 9535 Technical Manual GENERAL DESCRIPTION 1.2.2 CONTROL PANEL 14 13 12 11 10 9 8 1 2 3 4 5 6 7 Section 1 Doc. 9535 Page 9 of 65 Technical Manual GENERAL DESCRIPTION 1 2 kV kV increase mAs HP/LP (1) Power selection 4 Prog Unit ON 8 Unit OFF 9 Data storage in anatomical technique 10 Potter-Bucky selection 11 X-Ray exposure 12 X-Ray ready 13 Alarm signalling 14 Display for Dates and Messages mAs decrease mAs increase 3 7 kV decrease N° of exams memorized in Technique Anatomical 5 Anatomical Technique Selection 6 Collimator light activation (1) A change of the Power level involves a change of the used focal spot: • High Power (HP) – Big Focus • Low Power (LP) – Small Focus In the following paragraphs the abbreviations HP and LP will be used. Section 1 Page 10 of 65 Doc. 9535 Technical Manual GENERAL DESCRIPTION 2 TECHNICAL DATA 2.1 CLASSIFICATION OF THE DEVICE CLASSIFICATION – EN 60601 1 § 5 ➠ Type of protection against short circuit: CLASS I ➠ Degree of protection against direct and indirect contact: TIPE B ➠ Use conditions: CONTINUOUS WORKING WITH INTERMITTENT LOAD ➠ Unit not to be used in the presence of an inflammable anaesthetic mixture with air or nitrous oxide CLASSIFICATION – 93/42/EEC DIRECTIVE ➠ In according with Annex IX: CLASS II b 2.2 TECHNICAL CHARACTERISTICS Electrical characteristics 230 Vac (± 10%), 16 A 50/60 Hz SINGLE PHASE VOLTAGE FREQUENCY CONTINUOS WORKING 1A INTERMITTENT WORKING 12 A MAX ABSORBED CURRENT LINE COMPENSATION LINE RESISTANCE Automatic < 2.5 Ω Radiological characteristics 30 kW LP (Low Power) HP (High Power) MAX POWER 7.5 kW MAX CURRENT IN RADIOGRAPHY 150 mA EXPOSURE TIME WORKING FREQUENCY RANGE KV RANGE MAS A.T. PILOTAGE RIPPLE TOTAL FILTRATION RISING TIME 30 kW 425 mA 1.2 ms ÷ 1.0 s (chose from processor in function of mAs) 100 kHz 40 ÷ 125 (Step of 1 kV) 0.5 ÷ 200 in 25 values Inverter driven by IGBT ≤ 3% at Max Power > 2.7 mmAl ≤ 1 ms Section 1 Doc. 9535 Page 11 of 65 Technical Manual GENERAL DESCRIPTION X-Ray Tube Head All information relevant to the X-Ray Tube Head and to the X-ray tube can be found in the X-Ray Tube Head Technical Data Sheet Rotating with speed 3000 RPM TYPE OF ANODE 0.6 mm FOCAL SPOTS 1.3 mm Collimator (optional) All information relevant to the Collimator be found in the relative Technical Data Sheet SHUTTERS TO MULTIPLE PLANS Parallels and perpendicular with manual movement Dosimeter (optional) MODEL ACTIVE AREA MINIMAL DOSE RESOLUTION MAXIMAL MEASURABLE DOSE Kermax-plus 146 x 146 mm2 1 mGycm2 9999.9999 mGycm2 VacuDap 2000 with printer optional 147 x 147 mm2 1 mGycm2 9999.9999 mGycm2 Operating modes and functionality INTERFACE USER OPERATING MODES RADIOGRAPHY X-RAY CONTROL Polycarbonate flat keyboard with alphanumeric LCD display for all the operative parameters and messages of possible anomalous conditions – administrated by a microprocessor. Two-points techniques (kV-mAs) 40 programmable anatomic technique (20 for LP and 20 for HP) Distance control with double – click and extensible cable (≥4m) mAmin and mAmax Maximum exposure time Temperature maximum X-ray tube head SAFETY Count thermal units X-ray tube head Max kV, min kV, max ∆kV, max I Anode rotation Microprocessor self – test Transport and storage conditions MAXIMAL TEMPERATURE RECOMMENDED TEMPERATURE RELATIVE HUMIDITY ATMOSPHERIC PRESSURE –10°C ÷ 55°C 0°C ÷ 40°C 20% ÷ 90% 500 hPa ÷ 1060 hPa Operating conditions TEMPERATURE RELATIVE HUMIDITY ATMOSPHERIC PRESSURE 10°C ÷ 40°C 30% ÷ 75% 700 hPa ÷ 1060 hPa Section 1 Page 12 of 65 Doc. 9535 Technical Manual GENERAL DESCRIPTION Mechanical Characteristics WIDTH LENGHT HEIGHT WEIGHT 260 mm 498 mm 568 mm ≅ 90 kg NOTE: Upon your request it is possible to have a cable of 20m in order to position in a distance the control console and the double – click pushbutton (with extensible cable ≥4m). 2.2.1 DIMENSIONS AND WEIGHTS Drawing N°: 7465 Section 1 Doc. 9535 Page 13 of 65 Technical Manual GENERAL DESCRIPTION 2.2.2 kV – mA Relationship kV - mA Relationship in Radiogr aphy mode 30kW 470,00 420,00 mA Radiography 370,00 320,00 270,00 220,00 170,00 120,00 70,00 20,00 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 kV kV - mA Relationship in Radiogr aphy mode 7.5kW 160,00 mA Radiography 140,00 120,00 100,00 80,00 60,00 40,00 20,00 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 kV Section 1 Page 14 of 65 Doc. 9535 Technical Manual GENERAL DESCRIPTION 2.2.3 RX EXPOSURE TIME In the following tables the exposure times (s) associated to kV and mAs are reported. Since the apparatus is a device functioning with a two point technique (kV and mAs) we remind that the values indicated in the table are merely theoretical and can undergo a variation in relation to the tolerance of the mA. TABLE 1– 7.5KW POWER mAs 0,5 1 1,3 1,6 2 2,5 3,2 4 5 6,3 8 10 13 16 20 25 32 40 50 63 80 100 130 160 200 kV 40 0,003 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,042 0,053 0,066 0,086 0,106 0,133 0,166 0,213 0,266 0,333 0,420 0,533 0,666 0,866 1,066 1,333 (150mA) 50 0,003 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,042 0,053 0,066 0,086 0,106 0,133 0,166 0,213 0,266 0,333 0,420 0,533 0,666 0.866 1.066 (150mA) 60 0,004 0,008 0,010 0,012 0,016 0,020 0,025 0,032 0,040 0,050 0,064 0,080 0,104 0,128 0,160 0,200 0,256 0,320 0,400 0,504 0,640 0.800 1.040 (125mA) 70 0,004 0,009 0,012 0,014 0,018 0,023 0,029 0,037 0,046 0,058 0,074 0,093 0,121 0,149 0,186 0,233 0,299 0,373 0,467 0,588 0.747 0.934 (107mA) 80 0,005 0,010 0,013 0,017 0,021 0,026 0,034 0,043 0,053 0,067 0,086 0,107 0,139 0,172 0,215 0,268 0,344 0,430 0,537 0,677 0.860 1.075 (93mA) 90 0,006 0,012 0,015 0,019 0,024 0,030 0,038 0,048 0,060 0,075 0,096 0,120 0,156 0,192 0,240 0,301 0,385 0,481 0,602 0.759 0.963 (83mA) 100 0,006 0,013 0,017 0,021 0,026 0,033 0,042 0,053 0,066 0,084 0,106 0,133 0,173 0,213 0,266 0,333 0,426 0,533 0,666 0.840 1.066 (75mA) 110 0,007 0,014 0,019 0,023 0,029 0,036 0,047 0,058 0,073 0,092 0,117 0,147 0,191 0,235 0,294 0,367 0,470 0,588 0.735 0.926 (68mA) 120 0,008 0,016 0,020 0,025 0,032 0,040 0,051 0,064 0,080 0,101 0,129 0,161 0,209 0,258 0,322 0,403 0,516 0,645 0.806 1.016 (62mA) 125 0,008 0,017 0,022 0,027 0,034 0,042 0,054 0,068 0,085 0,107 0,136 0,169 0,220 0,271 0,339 0,424 0,542 0,678 0.847 1.067 (59mA) TABLE 2– 30KW POWER mAs 0,5 1 1,3 1,6 2 2,5 3,2 4 5 6,3 8 10 13 16 20 25 32 40 50 63 80 100 130 160 200 kV 40 0,001 0,002 0,003 0,003 0,004 0,005 0,007 0,009 0,011 0,014 0,018 0,023 0,030 0,037 0,047 0,058 0,075 0,094 0,117 0,148 0,188 0,235 0,305 0,376 0,470 (425mA) 50 0,001 0,002 0,003 0,004 0,005 0,006 0,008 0,010 0,012 0,015 0,020 0,025 0,032 0,040 0,050 0,062 0,080 0,100 0,125 0,157 0,200 0,250 0,325 0,400 (400mA) 60 0,001 0,002 0,003 0,004 0,005 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,034 0,042 0,053 0,066 0,085 0,106 0,133 0,168 0,213 0,266 0,346 (375mA) 70 0,001 0,002 0,003 0,004 0,005 0,007 0,009 0,011 0,014 0,017 0,022 0,028 0,036 0,045 0,056 0,070 0,090 0,112 0,140 0,177 0,225 0,281 (355mA) 80 0,001 0,002 0,003 0,004 0,005 0,007 0,009 0,011 0,014 0,018 0,023 0,029 0,038 0,047 0,059 0,074 0,094 0,118 0,148 0,186 0,237 0,296 (337mA) 90 0,001 0,003 0,004 0,005 0,006 0,007 0,010 0,012 0,015 0,019 0,025 0,031 0,040 0,050 0,062 0,078 0,100 0,125 0,156 0,197 0,25 (320mA) 100 0,001 0,003 0,004 0,005 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,043 0,053 0,066 0,083 0,106 0,133 0,166 0,21 0,266 (300mA) 110 0,002 0,004 0,006 0,007 0,009 0,012 0,015 0,019 0,024 0,030 0,039 0,049 0,063 0,078 0,098 0,122 0,156 0,196 0,245 0,308 (204mA) 120 0,003 0,007 0,009 0,011 0,014 0,017 0,022 0,028 0,035 0,044 0,056 0,070 0,092 0,113 0,141 0,177 0,226 0,283 0,354 0,446 (141mA) 125 0,004 0,008 0,010 0,013 0,016 0,020 0,026 0,032 0,040 0,050 0,063 0,079 0,103 0,127 0,159 0,198 0,254 0,317 0,397 0,500 (126mA) Section 1 Doc. 9535 Page 15 of 65 Technical Manual GENERAL DESCRIPTION Section 1 Page 16 of 65 Doc. 9535 Technical Manual Installation Section 2 Doc. 9535 Page 17 of 65 Technical Manual INSTALLATION 1 INSTALLATION 1.1 INTRODUCTION The installation procedure consists of different operations to be carried out in the order shown in the flow chart below. Electrical connections Acceptance Test The various operations indicated are described in detail in the following paragraphs. The installation process can only be considered completed after the TEST SHEET regarding the acceptance tests provided with this manual has been filled in. NOTE: the radiological system is normally pre-installed and configured in the factory according to the specific requirements of the customer. Certain adjustments may, however, be necessary when components are replaced. The whole machine parameter adjustment procedure is described in detail in this manual (SECTION 5). After 30 minutes of non-use, the equipment automatically switches into stand-by mode, that means the capacitors battery is unloaded; the words “Stand By” appears on the display of control console and the value of the battery voltage is ! zero. Pressing any key of control console, the battery reloads to the maximum value and it is possible to make X-Ray exposure. During capacitors battery loading the words “Waiting” will appear on control console display, as explained in SECTION 4 - § 2. Section 2 Page 18 of 65 Doc. 9535 Technical Manual INSTALLATION 1.2 ELECTRICAL CONNECTIONS The apparatus must be supplied with 230 V single-phase alternating power supply. The maximum absorbed current is reported on label of the unit and on technical data (SEE SECTION 1 - § 2.2). Make sure that the power supply socket is approved for 230 Vac – 16 A. ! 2 CONFIGURATION 2.1 CHANGING THE LANGUAGE FOR THE UNIT CONSOLE The signalling on the unit console can be displayed in Italian, English, French or Spanish. Selection of the language is made when the system is turned on as shown below. ➠ Turn the unit on using the ON/OFF button: the initial test of the unit will be activated automatically, during which the display will show the version of the control software installed (SEE SECTION 4 - § 2). ➠ Keep immediately the relevant button pressed until the word "WAITING" appears on the display in the language selected. ENGLISH ITALIAN FRENCH SPANISH Section 2 Doc. 9535 Page 19 of 65 Technical Manual INSTALLATION 2.2 PROGRAMMING THE ANATOMICAL TECHNIQUES Programming an ANATOMICAL TECHNIQUE means associating the KV and mAs values considered suitable with the identifier (i.e. with the name, such as "SKULL", "SHOULDER"…). It will then be possible to call up an ANATOMICAL TECHNIQUE by means of its name and immediately find the radiographic parameters already set to the values memorised previously. To associate the KV and mAs values with an identifier, proceed as follows. ➠ Press the ANATOMICAL TECHNIQUE SELECTION button: the LED in the corner of the button itself will light up ➠ using the PROGRAM SELECTION up-down type buttons, select the ANATOMICAL TECHNIQUE for which the KV and mAs values are to inserted; also select the power LP or HP where the values will be memorised ➠ select the KV and mAs values considered suitable for that type of examination using the KV SELECTION and mAs SELECTION buttons (if necessary see the SECTION 4 - § 3.2) ➠ Press the DATA MEMORISATION button until the acoustic signal is heard. At this point, the KV and mAs values relative to the selected ANATOMICAL TECHNIQUE have been memorised ➠ It is therefore possible to select a new ANATOMICAL TECHNIQUE to associate specific radiographic parameters to, by repeating the same procedure ➠ For exit to the programming the ANATOMICAL TECHNIQUES press the ANATOMICAL TECHNIQUE SELECTION button Section 2 Page 20 of 65 Doc. 9535 Technical Manual INSTALLATION 3 ACCEPTANCE TEST The ACCEPTANCE TEST includes all the control and checking operations listed on the TEST SHEET. These operations must be grouped together under two general categories: ➠ checking cables and connectors ➠ electrical checks Each operation is identified by a code which is then used on the TEST SHEET. 3.1 CHECKING CABLES AND CONNECTORS Visually control each cable and each connector, checking that there are no interruptions and/or crushing. A.1 Power supply cable of the unit A.2 Radiography control pushbutton cable A.3 Connection cable between unit/X-ray tube head 3.2 ELECTRICAL CHECKS With regard to the electrical part, correct operation of the following must be checked: ➠ SAFETY DEVICES ➠ SIGNALS NOTE: Remove the carters to carry out the electrical checks. The procedure is described below: B.1 POWER SUPPLY: connect the unit to the mains power supply; B.2 INITIAL TEST: check that the initial automatic test is correct of the unit; B.3 SOFTWARE: check that the visualization of installed software; B.4 LOADING: check that the switching on the unit appears the increment of voltage on display until 320 V. The complete loading of the capacitors battery is ready in about 40/45 sec; Section 2 Doc. 9535 Page 21 of 65 Technical Manual INSTALLATION B.5 DISCHARGING: switch the unit off and check the complete discharging of the capacitors battery by means the switching off the LD3 led mounted on S25 board; Do not carry out switching the unit On and Off ! B.6 SET-UP DATES IN MEMORY: enter on SET-UP (SEE SECTION 5 - § 3) and check that the SET-UP dates are in according with dates selected (SEE SECTION 5 - § 3 – TABLE 7); B.7 X-RAY TUBE FORMATION: when not in used the unit for a long time, before to use the unit normally, carry out the “formation” of the X-ray tube with a serial of exposure using the following dates (carry out for each value of kV and mAs 10 exposure and keep a pause of 1 minute between an exposure and another one): 40 kV 100 mAs 50 kV 80 mAs 60 kV 63 mAs 70 kV 63 mAs 80 kV 50 mAs 90 kV 50 mAs 100 kV 40 mAs 110 kV 40 mAs 120 kV 40 mAs During this formation check with oscilloscope the value of kV [S19/B board Tp11 (-); Tp3 (+) ] and mA [S19/B board: Tp11 (-); Tp6 (+) ] on inverter. Select the oscilloscope in the following way: CHANNEL A: 1 V/DIV ( 1V read is equal to 50 mA) CHANNEL B: 1 V/DIV ( 1V read is equal to 20 KV) BASE TIME: 25ms/DIV Section 2 Page 22 of 65 Doc. 9535 Technical Manual INSTALLATION In case of kV wave form presents of high voltage discharge or you hear discharge inside X-ray tube interrupt the sequence, wait few minutes before starting again; repeat the initial formation checking that the fuse F7 is interrupted (100A) (See. Electrical Draw); Warning: presence of 320 Vdc voltage! ! B.8 EXPOSURE TIME: check that during exposure time the X-ray passage led until the end of exposure the display shows the real exposure time in second (SEE SECTION 1 - § 2.2.3) B.9 STAND BY: let the equipment switched on in “Stand by” mode, that means without make X-Ray exposure, and verify that after 30 minutes the capacitors battery is unloaded, as explained in § 1.1. Press any key of control console to restore the capacitors loading and repeat the whole procedure for a total time of at least 4 hours. Section 2 Doc. 9535 Page 23 of 65 Technical Manual INSTALLATION Section 2 Page 24 of 65 Doc. 9535 Technical Manual Safety and Maintenance Section 3 Doc. 9535 Page 25 of 65 Technical Manual SAFETY AND MAINTENANCE 1 SAFETY 1.1 INTRODUCTION The aim of this manual is to provide qualified radiology technicians, medical and paramedical staff with operating instructions to make use of the RADIOGRAPHIC UNIT simple and safe. This apparatus emits X-RAYS and must only be used in compliance with the safety instructions indicated in this manual and must not be used for any other purposes than those foreseen. The system must only be used by personnel with the necessary knowledge in the field of radiation protection and with the necessary training in use of X-ray apparatus. We remind that the device is assembled inside a X-ray finished equipment, therefore the final user has to apply the unit safety indications. Some warnings are anyway hereby listed: ➠ the apparatus must not be used when there are any electrical and/or mechanical faults; ➠ do not use the system when any signalling or alarm device of the system is not working properly; ➠ under no circumstances work with the unit in places saturated with vapours and/or flammable gases or explosives; ➠ any modification to the system must be authorised in writing by the MANUFACTURER; ➠ if you want to use the apparatus in combination with other equipment, components or modules, and when compatibility with the latter is not certain, it is indispensable to make sure that there is no danger to the patients and/or operating personnel. In this case, consult the manufacturer of the apparatus in question or an expert in the sector; ➠ like any other apparatus, the System must be used correctly. It also need PERIODIC CHECKS AND MAINTENANCE as specified under § 2.1 in this section; ➠ all maintenance, repair and/or modification work must be carried out by PERSONNEL QUALIFIED AND AUTHORIZED by the Manufacturing Company. The latter declines any responsibility for malfunctions caused by unauthorised interventions; ➠ the MANUFACTURING COMPANY of the apparatus declines any responsibility for damage to people and/or things caused by its improper use. Section 3 Page 26 of 65 Doc. 9535 Technical Manual SAFETY AND MAINTENANCE 1.2 RESPONSIBILITY DECLARATION Enclosed to the present manual the MANUFACTURER RESPONSIBILITY DECLARATION (SEE ANNEX 1). 1.3 COMPLIANCE AND REFERENCE ADDRESS For information relevant to the compliance make always reference to ANNEX 1. 1.4 SYSTEM SAFETY 1.4.1 MECHANICAL SAFETY We remind that the device is assembled inside a X-ray finished equipment, therefore the final user has to apply the unit mechanical safety indications. 1.4.2 ELECTRICAL SAFETY ➠ Make sure that the power supply socket where the apparatus is to be connected is approved for the foreseen voltage and current for use of the system ➠ the radiology system must not be used in rooms where there is the risk of explosion ➠ disconnect the installation from the mains before carrying out any cleaning, disinfection and/or sterilisation ➠ always turn the apparatus off after use. Section 3 Doc. 9535 Page 27 of 65 Technical Manual SAFETY AND MAINTENANCE 1.5 PROTECTION AGAINST IONISING RADIATION Before carrying out any X-ray exposure, make sure that all the necessary precautions against radiation have been taken. During radiation emission, the personnel in the X-ray room must respect the regulations in force regarding protection against radiation. For this, bear in mind the following rules: ➠ where necessary, use protective accessories against radiation; ➠ always use the special X-ray protective coats: an X-ray protective material equivalent to 0.35 mm of lead (0.35 mm Pb) attenuates the radiation produced at 50 kV by 99.95% and at 100 kV by 94.5%; ➠ the best protection against radiation is distance: therefore keep as far away as possible both from the source of radiation and from the exposure object, also by using the suited exposure-push-button with its extensible cable; ➠ avoid moving or standing in the path of the rays; ➠ always use the smallest exposure range possible: the radiation dispersed largely depends on the volume of the object X-rayed; ➠ keep the patient’s examination area the furthest away as possible from X-ray source. Section 3 Page 28 of 65 Doc. 9535 Technical Manual SAFETY AND MAINTENANCE 1.6 RESIDUAL RISKS The system is designed and constructed according to the strictest principles of compliance with safety requirements. However, there are residual risks due either to incorrect use of the apparatus or to deficiencies in the protective measures taken. With regard to the risks due to incorrect use of the apparatus, please see the instructions and recommendations given in the points above. For the residual risks due to any defect in the protection measures taken, it must be remembered that: ➠ Protection against electric shocks is carried out by means connecting the metallic covering parts of the apparatus to earth: it is therefore necessary to periodically check – according to the NORMAL MAINTENANCE PLAN described under § 2.1 of this section – correct operation of the whole earthing circuit. Not taking notice of the unit alarms could cause overheating of the X-RAY TUBE ! HEAD: this overload could lead to loss of the means of insulation in the X-ray tube head itself at very high temperatures. Section 3 Doc. 9535 Page 29 of 65 Technical Manual SAFETY AND MAINTENANCE 1.7 SIGNALS 1.7.1 SYMBOLS USED Apart from the symbol on the CONTROL CONSOLE, others have been used on the unit, as illustrated below. SYMBOL MEANING POSITIONING CAUTION: SEE THE ATTACHED DOCUMENTATION Unit label, protection panel of the electronic group PHYSIOLOGICAL EFFECTS Unit label, control panel CLASS B EQUIPMENT (IEC-601-1 STANDARD) Unit label EQUIPMENT THAT REQUIRES A CORRECT DISPOSAL (2002/96/EC) Unit label PROTECTIVE EARTH Equipotential bar RADIOGRAPHY Radiography pushbutton X-RAY FOCUS POSITION X-ray tube head covering 1.7.2 LABELLING Sample of the IDENTIFYING LABELS sticked on the X-ray unit is attached to the present manual (SEE ANNEX 2). Section 3 Page 30 of 65 Doc. 9535 Technical Manual SAFETY AND MAINTENANCE 2 MAINTENANCE 2.1 ROUTINE MAINTENANCE 2.1.1 GENERAL RACOMMENDATIONS The radiological system requires regular checks and maintenance. The following recommendations have the aim of helping the operator to keep the apparatus in good working and safe conditions during service. Correct adjustment of the electrical and electronic systems has a direct influence on the operation of the system, on the quality of the image and on the electrical safety of the system, as well as on the level of exposure to radiation the operators and patients are subjected to. The MAINTENANCE PROGRAMME, described in the following paragraphs, consists of controls and interventions to be carried out by specialised personnel authorised by the manufacturer. All maintenance operations are the responsibility of the owner of the apparatus. Should it be necessary to replace components or parts which may in any way ! condition the safety of the machine, only use original spare parts. Section 3 Doc. 9535 Page 31 of 65 Technical Manual SAFETY AND MAINTENANCE 2.1.2 FREQUENT CHECKS AND INSPECTIONS The operating personnel must be suitably trained to be able to carry out the daily and weekly checks indicated in TABLE 3. The other controls described in this chapter and the interventions described in the following chapters are reserved for qualified and authorised personnel of the technical assistance service. TABLE 3 INTERVAL CHECK DAILY Check operation of the signals, displays and LEDs Check that the warning and danger labels are intact CHECKS Check that there are no oil leaks from the X-ray Tube Head Check there is no unusual noise in the X-ray Tube Head during X-ray emission WEEKLY CHECKS 6-MONTHLY CHECKS Check the correct operation and the value of the whole earthing circuit. Check the power supply voltage value Check the value of the continuous voltages generated inside the system Check fixing and general state (dust and corrosion) of the boards 2.2 GENERAL CHECKS AND INSPECTIONS Every six months, and in any case according to local legislation in force, the procedures indicated in the ACCEPTANCE TEST REPORT must be carried out completely. The results obtained are noted down on the TEST SHEET together with the description of any interventions carried out. The checks listed below must be added to those contained on the Test Sheet: Power supply voltage check DC power supply check Protection earth check Internal earth connection check 2.3 CLEANING AND DISINFECTION The end user has to apply. Section 3 Page 32 of 65 Doc. 9535 Technical Manual SAFETY AND MAINTENANCE 2.4 SPECIAL MAINTENANCE 2.4.1 TROUBLESHOOTING Troubleshooting can be facilitated by the indications given in TABLE 4, which illustrates the most common incorrect operating conditions, showing their probable causes. In the presence of the following alarms visualized on the display, the X-ray exposition is disqualified and the console’s alarm red led light up. In case it is also present after the execution of the suggested intervention, please contact the service assistance department. TABLE 4 N° FAULT NOTED LIKELY CAUSE RECOMMENDED INTERVENTION The unit doesn’t switch on (on key LED No power supply on input Check / Replace the S22-F2 or TR2switched on) F24V fuse The unit doesn’t switch on (on key LED No power supply on input Check / Replace the S22-F1 or F1 switched off) fuse (on input) Check the Radiography pushbutton 2 The apparatus works but does not emit Radiography pushbutton fault rays – No alarm on the unit display cable and connector +15V or –15V fault of the S22 Check power supply (+15VA/3 Alarm: “SUPPLY FAULT” board supply 15VA/+24VA voltage on S22 voltage) 400 Hz power supply absent Check / Replace TR2-F130V0 fuse 4 Alarm: “FILAMENT FAULT” X-ray tube filament interrupted Replace the X-ray tube head S15 board fault Replace the board If the indication appears on stand- Carry out set-up as in Section 5 - § 5 Alarm: “INVERTER FAULT” by: P1 Set-Up on S22 board faulty 4.2 If the indication appears while Switching off the unit, switching it on carrying out X-ray: faulty on X-ray and repeat the X-ray power circuit F2 and F3 fuses blown Replace F2 and F3 fuses F6 and F7 fuses blown on the Replace F6 and F7 fuses Power circuit During a radiograph the effective Switching off the unit, switching it on 6 Alarm: “KV FAULT” kV are less than 85% of those set: and repeat the X-ray fault on the power circuit During an exposure the mA value Switching off the unit, switching it on 7 Alarm: “mA FAULT” is lower than the allowed limit and repeat the X-ray X-ray tube head overheated Do not carry out any X-rays and wait 8 Alarm: “THERMIC SAFETY” for the X-ray tube head to cool down Breakage of the thermal sensor Replace the thermal sensor located on the X-ray tube head and/or check the relevant wiring on S22 board During a radiograph with cassette, Repeat the radiography 9 Alarm: “MAN STOP RX” the control pushbutton for X-ray command has been released early Error in the high voltage generation Switching off the unit, switching it on 10 Alarm: “X-RAY LACKING” circuit and repeat exposure 1 Section 3 Doc. 9535 Page 33 of 65 Technical Manual SAFETY AND MAINTENANCE N° FAULT NOTED LIKELY CAUSE 11 Alarm: “MAX TIME” RECOMMENDED INTERVENTION 14 Alarm: “BATTERY FAULT” The radiography time has exceeded Check that the mA are correct the concurred limit Check the mA reading on board S22Tp1 or S19/B-Tp6 During anode preparation it was Check the fuses on board S21ACheck not put into rotation the stator winding of the X-ray tube head Fault on board S21/A Replace board S21/A Fault on board S20 Replace board S20 Capacitors bank still loads With the unit off wait 1 minute before switching it on Power Circuit Fault Contact technical service 15 Alarm: “OVERVOLTAGE BATTERY” Battery Circuit Fault 16 Alarm: “X-RAY COMMAND ACTIVE” The operator has pressed the Release the radiography pushbutton radiography command before the and wait until the system is ready system had finished the initial control stage It is not possible to begin exposure Wait for the X-ray tube to cool down. since the remaining thermal units available are too few The keyboard does not Switching off the unit, switching it on communicate with the unit and repeat exposure Potter Fault Check the F4 - F5 fuses of the Potter power supply and the K3 relè 12 Alarm: “STARTER FAULT” 13 Alarm: “SWITCH OFF FOR 1 MIN” 17 Alarm: “X-RAY TUBE TOO HOT” 18 Alarm: “WAIT CONNECTION” 19 Alarm: “POTTER FAULT” (only with Potter installed) Contact technical service 2.4.2 REPLACEMENT OF COMPONENTS Replacement of a component can mean the need to carry out checks and adjustments in order to reset correct operation of the radiological unit. TABLE 5 shows typical situations where an adjustment is necessary. TABLE 5 COMPONENT REPLACED 15 BOARD S19/B BOARD S20 BOARD ADJUSTMENT AND CHECKS S25 BOARD Verify mA set up in SET UP Check of kV e mA Check Run voltage Check start 800 ms Check max Set kV (P1) Check max exposure time (P2) Check max voltage of battery S62 BOARD Check LCD Display contrast (P1) S22 BOARD REFERENCE SECTION 5 - § 5 SECTION 5 - § 5.1 and 8 SECTION 5 - § 2 SECTION 5 - § 4.3 SECTION 5 - § 4.1 and 4.2 SECTION 5 - § 7 Section 3 Page 34 of 65 Doc. 9535 Technical Manual SAFETY AND MAINTENANCE 2.4.3 LIST POTENTIOMETERS BOARD POTENTIOMETER VALUE NAME NOTE S15 5K 5K 5K 10K 10K 10K 10K 2K 2K 2K MANUAL SET I FIL (ONLY FOR TEST) MAX I FG MAX I FP S25 P1 P2 P3 P1 P2 P3 P4 P1 P2 P3 R66 P1 P2 P1 P2 P1 1K 1M 2K 100K 5K MAX CURRENT FREQUENCY SET MA FLUORO OFFSET ADJ. FEEDBACK MA ADJ RUN VOLTAGE START 800MS MAX SET KV MAX TIME RAD MAX V CAPACITORS BANK Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment N.U. Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment Only factory adjustment S62 P1 20K LCD CONTRAST Only factory adjustment S17 S19/B S20 S22 FEEDBACK KV+ FEEDBACK KV- Section 3 Doc. 9535 Page 35 of 65 Technical Manual SAFETY AND MAINTENANCE 2.4.4 LIST LEDS BOARD LED COLOR NAME S15 LD1 LD2 LD3 LD1 LD2 LD3 LD4 LD5 LD6 LD7 LD8 LD9 LD10 LD1 LD2 LD3 DL1 DL2 DL3 DL4 DL5 DL6 DL7 DL8 DL9 DL10 DL11 DL12 DL13 DL14 DL15 DL16 DL17 DL18 DL19 DL20 DL21 DL22 DL23 DL25 DL26 DL27 DL28 DL29 DL30 DL31 YELLOW GREEN RED YELLOW RED RED RED RED YELLOW YELLOW YELLOW GREEN GREEN YELLOW GREEN YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW GREEN YELLOW GREEN YELLOW YELLOW YELLOW YELLOW GREEN GREEN YELLOW GREEN YELLOW YELLOW YELLOW YELLOW YELLOW YELLOW +24V +15VA SAFETY MAX I FIL COM-RX ∆ KV MAX KV > 110% KV MIN I MAX COM. 1 COM. 2 PRE-RX +15V -15V STARTER ON STARTER OK FLUORO PREP (N.U.) N.U. N.U. N.U. N.U. N.U. N.U. N.U. N.U. STARTER OK PREP K1A RAD KV 85% POTTER OK THERMIC KV OK FIL OK PRE RX RX PREP K1A START POTTER +24VA +24VA ON/OFF +24VA CAPACITORS BANK CHANGE CAPACITORS BANK CHARGE OUT_WD PREP STARTER FP CORR MA S19/B S20 S22 NOTE Section 3 Page 36 of 65 Doc. 9535 Technical Manual SAFETY AND MAINTENANCE 2.4.5 LIST FUSES BOARD NAME TYPE LOADING BATTERY F2 F3 F6 F7 F1 F230V F115V F135V F24V F12V F4V F5V F1 F2 F3 F1 F2 F1 F1 F2 F3 F4 F5 F6 F7 F8 F9 16A 16A 63A 63A 200MA-T 2A-T 3.15A-T 1A-T 2A-T 10A-T 2A-T 2A-T 10A-T 10A-T 500MA-T 1A-T 2A-T 80mA-T 200mA-T 200mA-T 200mA-T 200mA-T 200mA-T 200mA-T 200mA-T 200mA-T 315mA-T INVERTER TR1 TRANSFORMER TR2 TRANSFORMER POTTER BUCKY S21/A S22 S25 S29 NOTE Input power supply protection Protection Inverter Protection TR1 Transformer Protection TR2 Transformer Protection Potter Bucky (Optional) Input power supply protection S21/A Board Protection TR1 Transformer S21/A Board Protection Bridge Rectifier (24Vac) Input power supply protection S25 Board Protection Bridge Rectifier Group 1 (24Vac) Protection Bridge Rectifier Group 2 (24Vac) Protection Bridge Rectifier Group 3 (24Vac) Protection Bridge Rectifier Group 4 (24Vac) Input power supply protection S29 Board Section 3 Doc. 9535 Page 37 of 65 Technical Manual SAFETY AND MAINTENANCE 2.5 SPARE PARTS In case of replacing of parts that can be negative for the safety of the machine, use only original parts. For the list of the spare parts and relative codes to make reference to the SECTION 6 – EXPLODED DRAWINGS. NOTE: When you require spare parts ,it is necessary to tell the code of the piece and the serial number of the unit too. 2.6 DISPOSAL OF DEVICE Once the product is at the end of its use it must be disposed following the enforced rules concerning the separate collection of waste and it cannot be treated as a simple urban waste. The symbol indicates that the product has the requirements requested by new directives introduced for the environmental protection (2002/95/EC, 2002/96/EC, 2003/108/EC) and it must be disposed properly once its life cycle is ended. When the device has reached the end of its life it must be disposed at the proper centres for the separate collection of electrical and electrical wastes, or it must be returned to the reseller or to the manufacturer in case you want to replace the product with another equivalent new one. The proper separate collection helps to avoid possible negative effect on the environment and on health and it facilitates the second use and the re-cycle of materials of which the device is composed of. Ask for further information to the local authorities about the areas dedicated to the wastes disposal. Who does not dispose the product following what here above mentioned will be responsible for the enforced rules. Section 3 Page 38 of 65 Doc. 9535 Technical Manual Use of the unit Section 4 Doc. 9535 Page 39 of 65 Technical Manual USE OF THE UNIT 1 TRANSPORT We remind that the device is assembled inside a X-ray finished equipment, therefore the final user has to apply the indications about the transport of the unit. 2 SWITCHING ON To connect the unit to the power supply mains, only use 16A sockets with ! earthing terminal. ! Do not carry out continuous switching on and off of the unit. To make the system operational, follow the indications given below: ➠ connect the unit to the power supply mains: on unit the LED of the control console remain lit ➠ turn the unit on ➠ the initial automatic Test of the unit will be activated immediately: • for about 4 seconds, the display shows the version of control software installed; all the leds light up in sequence 30 E100 R. X.XX UNIT CONTROL CONSOLE – MESSAGES SHOWN ON SWITCHING ON Section 4 Page 40 of 65 Doc. 9535 Technical Manual USE OF THE UNIT ➠ successively will appear on the display the message WAITING until the capacitors battery loading are ready, the value of the battery voltage and the job technique. NOTE: Before carrying out any other operation, wait until the wording “WAITING” disappears from the display. WAITING 250V MAN UNIT CONTROL CONSOLE – MESSAGES SHOWN ON SWITCHING ON ➠ the display shows the preset radiography values and the working technic (technic preset manual), indicates the percentage of thermal units still available (when the X-ray tube head is under starting conditions, the display will indicate a percentage of H.U. of 100%) ➠ in this situation the machine is ready for X-ray command ➠ select the type of exposure required according to the description on the following pages. Section 4 Doc. 9535 Page 41 of 65 Technical Manual USE OF THE UNIT 40 0.5 100% MAN UNIT CONTROL CONSOLE – MESSAGES SHOWN WHEN SYSTEM IS READY Section 4 Page 42 of 65 Doc. 9535 Technical Manual USE OF THE UNIT 3 EMISSION OF X-RAYS Before carrying out any exposure, make sure that all the necessary precautions ! against radiation have been taken. Whatever the exposure mode chosen, X-ray passage is signalled ! by the LED lighting up. After 30 minutes of non-use, the equipment automatically switches into stand-by mode, that means the capacitors battery is unloaded; the words “Stand By” appears on the display of control console and the value of the battery voltage is ! zero. Pressing any key of control console, the battery reloads to the maximum value and it is possible to make X-Ray exposure. During capacitors battery loading the words “Waiting” will appear on control console display, as explained in § 2 of this Section. STAND BY . . . 0V MAN Section 4 Doc. 9535 Page 43 of 65 Technical Manual USE OF THE UNIT 3.1 EMISSION OF X-RAYS USING THE MANUAL TECHNIQUE ➠ After being switched on (SEE § 2 of this section), the unit automatically goes into MANUAL TECHNIQUE on Low Power LP; if you like select High Power HP use the button ➠ the preset KV and mAs values appear on the display ➠ using the KV SELECTION and mAs SELECTION up-down buttons, set the required values for these parameters (1) ➠ pull out the X-ray Control Push Button. This device is a doubleclick button (Click A and Click B – see figure) and must be used as indicated below A B ➠ use the X-RAY CONTROL pushbutton and make the first click (A); keep the button in this position until on the display appears the wordings READY FOR RADIOGRAPHY and the X-ray ready LED lights up ➠ make the second click (B) with the X-RAY CONTROL pushbutton and keep it in that position. At that moment X-ray emission starts and the X-ray on LED lights up Section 4 Page 44 of 65 Doc. 9535 Technical Manual USE OF THE UNIT ➠ at the end of X-ray emission, the X-ray ON LED turns off and the acoustic signal stops. At this point both of the X-ray control clicks can be released and the display shows the actual emission time. Releasing the first click (A) too soon has no effect. Releasing the second click (B) too soon interrupts X-ray emission; the display will ! show the message “MAN STOP RX” for about 10 seconds and the unit will give an alarm sound. Section 4 Doc. 9535 Page 45 of 65 Technical Manual USE OF THE UNIT 3.2 X-RAY EMISSION USING THE ANATOMICAL TECHNIQUE The X-ray unit allows up to 20 anatomical techniques to be memorized for each power (LP and HP). ANATOMIC TECHNIQUE means preset KV and mAs values which have been assigned a more easily understood name for the user. The identifiers (i.e. the names) of each of the 20 Anatomical techniques which can be programmed on the unit are shown in table. Programmable anatomical techniques N° Exam KV mAs 1 SKULL 75 40 2 CERVICAL 65 16 3 BACK-BONE 90 50 4 LUMBAR 95 50 5 LUNG 110 1 6 SHOULDER 70 20 7 COLLAR-BONE 70 10 8 HUMERUS 50 10 9 ELBOW 48 10 10 INFERIOR ARM 46 10 11 W RIST 42 10 12 FINGER 40 5 13 PELVIS 75 20 14 FEMORAL JOINT 70 25 15 ABDOMEN 85 16 16 FEMUR 70 16 17 KNEE 60 6.3 18 INFERIOR LEG 55 10 19 ANKLE 55 8 20 FOOT 48 6.3 Section 4 Page 46 of 65 Doc. 9535 Technical Manual USE OF THE UNIT It is possible to select one of the ANATOMICAL TECHNIQUES programmed previously at any time and to use the memorized radiographic parameters. This procedure is described in the following steps: ➠ select the power and consequently the filament which you desire to use ➠ press the ANATOMICAL TECHNIQUE TELECTION: the LED in the corner of the button itself will light up ➠ using the PROGRAMME SELECTION up-down type buttons, select the anatomical technique to be used NOTE: to make a radiography is possible to use the stored dates or change the value using the KV SELECTION and mAs SELECTION up-down buttons. If you desire to memory the variation of the value compare with SECTION 2 - § 2.2; differently, when you go out from the modality “Anatomical technique” the value variation will be lost. ➠ make the exposure as described in § 3.1 X-RAY EMISSION IN MANUAL TECHNIQUE starting from point (1). 4 OPERATIONS AFTER USE We remind that the device is assembled inside a X-ray finished equipment, therefore the final user has to apply the indications about the operations to effect once ended to use it. Section 4 Doc. 9535 Page 47 of 65 Technical Manual USE OF THE UNIT Section 4 Page 48 of 65 Doc. 9535 Technical Manual X-ray settings Section 5 Doc. 9535 Page 49 of 65 Technical Manual X-RAY SETTINGS 1 INTRODUCTION The radiological system is fully regulated in the factory: it is therefore only necessary to install it in order to use it. All the information contained in this section and SECTION 3 of this manual must be used when extraordinary maintenance is necessary. In the case of any intervention, it is essential for the adjustment operations to be ! carried out in the sequence indicated. In TABLE 6 indicates the potentiometers which do not require any further ! adjustments since they have already been regulated and sealed in the factory. TABLE 6 BOARD POTENTIOMETER DESCRIPTION S15 P1 P2 P3 P1 P2 P3 P4 P1 P2 P1 P2 P1 Manual Setting of the filament current Maximum current of large focus Maximum current of small focus S17 (X-ray tube head) S19/B S22 S25 Return of the kV+ Return of the kVMaximum current Frequency setting Maximum kV setting Maximum exposure time Maximum battery voltage NOTE : Jumper S19/B-J2A and J2B ON: safety kVmax set up for rotating anode Jumper S19/B-J2A and J2B OFF: safety kVmax set up for stationary anode Section 5 Page 50 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS 1.1 DIP SWITCH Here follows the function of the dip-switch S22-SW1 pins: 1 – Only factory adjustment 6 – Only factory adjustment 2 – Only factory adjustment 7 – Only factory adjustment 3 – ON – It doesn’t display the name when switched on 8 – Only factory adjustment 4 – Only factory adjustment 5 – ON – Small Focus when switched on 2 ROTATING ANODE STATOR POWER SUPPLY ADJUSTMENT ➠ Connect an AC voltmeter (true RMS if possible) with full-scale set fs: 300V at 300 V between the TB3-2 and TB3-3 test points of board S21/A (power supply of the main winding); TB3-2 TB3-3 ➠ connect the probe of an oscilloscope between IC2-8 (+) and Dz1 Anode (-) of board S20 and prepare the instrument to read a time of 800 ms; ➠ take the Radiography Command pushbutton. For maximum operator safety, at this point it is advisable to temporarily disable ! the X-ray command signal by disconnecting the wire coming from S22-X4-3. ➠ only press the first click of the Radiography Command pushbutton (preparation pushbutton); ➠ check that the stator is supplied with a voltage of about 230 Vac for approximately 0.8 seconds. If necessary, adjust this time interval using potentiometer P2 of board S20; ➠ check that once the 0.8 seconds have passed the stator is supplied with a voltage of about 70 Vac. If necessary, adjust this value by using poten-tiometer P1 of board S20. Section 5 Doc. 9535 Page 51 of 65 Technical Manual X-RAY SETTINGS 3 SET-UP PROCEDURE To access the modality of SET-UP procedure to follow what underneath reported: ➠ press the keys of the console indicated to side at the same time ➠ the wording "SET-UP PROCEDURE ?" will appear on the display; YES to confirm selection, press the key indicated at the side marked with "YES", whereas to cancel the procedure, press the key NO marked "NO". NOTE: In the Set-up procedure, the functions for editing the content of the fields shown on the display have been assigned to the keys as indicated below. YES + kV mAs NO - TABLE 7 – MINIMUM, MAXIMUM AND OF DEFAULT VALUES OF SET-UP ADJUSTMENT FUNCTION DIRECTORY VALUE FILE MINIMUM MAXIMUM DEFAULT KV40 300 600 380 kV120 1000 1500 1190 150 mA (40 kV) 1000 2000 1600 75 mA (100 kV) 900 2000 1360 425 mA (40 kV) 1000 2000 1750 300 mA (100 kV) 900 2000 1560 MAS ADJ kV 050 (50 mAs) 250 1000 530 BATTERY V = 320 kV ADJUSTMENT SMALL FIL. ADJ LARGE FIL. ADJ 7.5 kW 30 kW Section 5 Page 52 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS 3.1 SERVICE MENU When the unit is off, pushing both the keys underneath you enter the SERVICE MENU of the unit which subsequently visualizes the following data: ➠ name of the device and software version ➠ chosen anatomical program ➠ X-ray tube model ➠ total number of radiographic exposures ➠ total mAs accumulated. k mA V s Section 5 Doc. 9535 Page 53 of 65 Technical Manual X-RAY SETTINGS 4 KV ADJUSTMENT 4.1 ADJUSTMENT “SET-UP” VOLTAGE The SET KV voltage is supplied by the microcomputer and must be adjusted according to the relationship kV/20 (i.e.: 1V = 20kV). ➠ First of all, enter the SET-UP procedure, following the instructions indicated in the previous paragraph; ➠ On board S19/B connect the probes of a Vdc digital multimeter between CP1-10 (+) and TP11 (-); ➠ in Set-up select "kV ADJ at 40 kV"; kV ADJ a 40 kV 2,00 V ➠ adjust the value until a reading of 2,00 V is obtained on the + - CP1-10 TP11 voltmeter; ➠ in SET-UP select "kV ADJ at 120 kV"; kV ADJ a 120 kV 6,00 V ➠ adjust the value until a reading of 6,00 V is obtained on the voltmeter; ➠ exit the SET-UP procedure by using the key; + - CP1-10 TP11 ➠ then select 100 kV, 80 kV and 60 kV, checking that the voltage read on the voltmeter is Vread = (KvSET/20) ± 25 mV. 4.2 ADJUSTMENT OF MAXIMUM KV SETTING SAFETY The maximum parameter value of SET KV under operating conditions must be 6 V, corresponding to 120 kV. The MAX SET KV safety device intervention is normally set to 130 KV. ➠ connect the probes of a Vdc digital voltmeter between test points TP10 (+) and TP9 (-) of board S22; ➠ turn the system on and use potentiometer P1 until a reading of 6,5 V ± 0,1 V is obtained. 6,5 V + - TP10 TP9 Section 5 Page 54 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS 4.3 CHECKING MAXIMUM KV SETTING SAFETY ➠ Connect the voltmeter to test points TP10 (+) and TP9 (-) of board S22 ➠ lower the MAX SET KV safety intervention to 100 kV adjusting potentiometer P1 until a reading of 5 V ± 0,1 V is obtained on the voltmeter; 5V + - TP10 TP9 ➠ checking that by setting the kV to a higher value than 100 kV the "KV FAULT" alarm goes off; 6,5 V ➠ put the MAX SET KV safety intervention back to 130 kV using potentiometer P1 again until a reading of 6.5 kV ± 0,1 V is obtained. + - TP10 TP9 Section 5 Doc. 9535 Page 55 of 65 Technical Manual X-RAY SETTINGS 5 ADJUSTMENT OF THE FILAMENT CURRENT The following tables indicate the mA anode values associated with the kV and the typical filament currents determined on a sample X-ray tube head for 30 kW and 7.5 kW power. TABLE 8 RADIOGRAPHY – 30 KW Set kV kV mA IFilament S22: TP10 – TP9 on display S19/B: TP6 - TP11 S15: CP6-2 [V] [kV] [1V=10mA] [mA] 2 40 425 870 2,5 50 400 817 3 60 375 822 3,5 70 355 803 4 80 337 814 4,5 90 320 806 5 100 300 795 5,5 110 204 769 6 120 141 753 Pre-switching: 466 TABLE 9 RADIOGRAPHY – 7.5 KW Set kV kV mA IFilament S22: TP10 – TP9 on display S19/B: TP6 - TP11 S15: CP6-2 [V] [kV] [1V=10mA] [mA] 2 40 150 826 2,5 50 150 777 3 60 125 757 3,5 70 107 747 4 80 93 737 4,5 90 83 727 5 100 75 716 5,5 110 68 706 6 120 62 696 Pre-switching: 471 Section 5 Page 56 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS 5.1 ADJUSTMENT OF FILAMENT CURRENT NOTE: by Filament Current the current which runs through the primary winding of the transformer supplying the power to the filament itself is intended (Coolidge Transformer). ➠ MEASUREMENTS • To measure the Filament Current, connect an ammeter with true fs: 1 A RMS reading in series with terminal CP6-2 of board S15. Set the full-scale to 1A in AC. • to measure the anode mA in RADIOGRAPHY, connect the probe CP6 1V=50mA of an oscilloscope between test points TP6 (+) and TP11 (-) of + - board S19/B to display the wave form of the mA (reading TP6 TP11 correspondence is 1V = 50 mA) and between test points TP3 (+) and TP11 (-) of board S19/B to display the wave form of the kV (reading correspondence is 1V = 20 kV). 1V=20kV + - TP3 TP11 ➠ PROCEDURE (example with Small focus LP 7,5 kW) • activate the SET-UP procedure; • select "SMALL FIL ADJ"; • adjust the "DAC" value set until a filament current, displayed on the digital voltmeter, equal to the one indicated in TABLE 9 for 40 kV (826) is obtained – Typical DAC 1600 – FIGURE 1; • give the X-ray command and check that the anode mA are 150. If necessary, adjust the SET-UP until the desired value is obtained; • proceed with SET-UP; • adjust the set value of the "DAC" until a filament current, displayed on the digital voltmeter, equal to the one indicated in TABLE 9 for 100 kV (716) is obtained – Typical DAC 1360 – FIGURE 2; Section 5 Doc. 9535 Page 57 of 65 Technical Manual X-RAY SETTINGS • give the X-ray command and check that the anode mA are 75; if necessary, adjust the SET-UP until the desired value is obtained; • proceed with SET-UP by selecting "LARGE FIL ADJ"; • Repeat the same operations described for "SMALL FIL ADJ" with 40 kV and 100 kV. For the anode and filament current mA values, refer to TABLE 8. • Exit SET-UP. 826 mA 150mA (40kV) 1600 FIGURE 1 – DAC (40 KV) ADJUSTMENT 716 mA 75mA (100kV) 1360 FIGURE 2 – DAC (100 KV) ADJUSTMENT Section 5 Page 58 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS 6 mAS ADJUSTMENT ➠ MEASUREMENTS • Remove the bridge between terminals CP3-1 and CP3-3 of board S19/B • connect a mAsmeter between the above-mentioned terminals. ➠ PROCEDURE • activate the SET-UP procedure • select "mAS ADJ" and give the X-ray command • check that the mAs value on the instrument is really the one set (50 mAs – Typical DAC 0530); If necessary, adjust the SET-UP until the desired value is obtained 50 mAs + - CP3-1 CP3-3 • exit SET-UP. Section 5 Doc. 9535 Page 59 of 65 Technical Manual X-RAY SETTINGS 7 ADJUSTMENT OF THE BATTERY VOLTAGE ➠ MEASUREMENTS • To measure the battery voltage, connect a voltmeter with true RMS reading between terminals CP2-1 (+) and CP2-3 (-) of board S25. 320 V + - CP2-1 CP2-3 Warning: presence of 320 Vdc voltage ! ➠ PROCEDURE • activate the SET-UP procedure; • select "V BATTERI ADJ"; • Adjust the Set-up until the point in which the battery voltage measure value, reaches the battery voltage set value (tip. 320V); • exit SET-UP. Section 5 Page 60 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS 8 CHECKING THE RADIOGRAFIC PARAMETER After having completed the operations indicated in the previous paragraphs (§ 2, § 4, § 5), it is necessary to check the values of the signals proportional to the kV and to the mA which come from the INVERTER. The tolerance for the kV signal is ± 5% whereas for the mA it is ± 15%. NOTE: If the kV value is not correct, it is necessary to repeat the SET kV adjustment. NOTE: An oscilloscope with two channels is required for the checks below. 8.1 CHECKING KV AND mAs DURING RADIOGRAPHY ➠ Connect the probe of the oscilloscope as indicated below: Ch. A CHANNEL A: TP3 (+) – TP11 (-) of board S19/B (REAL kV); TIME BASE: 10 ms + - TP3 TP11 ➠ If necessary, connect the probe of the oscilloscope as indicated Ch. B below: CHANNEL B: TP6 (+) – TP11 (-) of board S19/B (mA); + - TP6 TP11 TIME BASE: 10 ms ➠ set the minimum value of the kV (40 kV – 425 mA) ➠ give the X-RAY command and check that the values read on the oscilloscope correspond with those indicated in TABLE 10; (FIGURE 3 shows an example of a 100 kV wave form) ➠ repeat the previous operations, setting the kV values to 80 kV and 120 kV for 7.5 kW power. TABLE 10 – 30KW kV mAs mA Time (s) Set Acceptance range S19: TP3 - TP11 [1V = 20 kV] Set Acceptance range (±10%) 40 38 ÷ 42 16 14.5 ÷ 17.5 425 0.037 80 76 ÷ 84 16 14.5 ÷ 17.5 337 0.047 120 114 ÷ 126 16 14.5 ÷ 17.5 141 0.113 Theoretical Theoretical Section 5 Doc. 9535 Page 61 of 65 Technical Manual X-RAY SETTINGS TABLE 11 – 7.5KW kV mAs mA Time (s) Set Acceptance range S19: TP3 - TP11 [1V = 20 kV] Set Range di accettazione (±10%) 40 38 ÷ 42 8 7.2 ÷ 8.8 150 0.053 80 76 ÷ 84 8 7.2 ÷ 8.8 93 0.086 120 114 ÷ 126 8 7.2 ÷ 8.8 62 0.129 Theoretical Theoretical FIGURE 3 – ACQUISITION OF THE KV WAVE FORMS DURING RADIOGRAPHY [100 KV – 16 MAS – 30 KW] FIGURE 4 – ACQUISITION OF THE MA WAVE FORMS DURING RADIOGRAPHY [100 KV – 16 MAS – 30 KW] Section 5 Page 62 of 65 Doc. 9535 Technical Manual Drawings Section 6 Doc. 9535 Page 63 of 65 Technical Manual DRAWINGS Index Document n° Electrical diagrams Electronic group 100 8795 S15 board – filament power supply 7616 S17 board – X-ray tube head interface 7548 S19/B board – inverter control 7617 S20 board - X-ray tube stator starter 7618 S21/A board – X-ray tube stator rotation 7619 S22 board – microprocessor 7620 S25 board – capacitors bank control 7621 S62/A board – keyboard interface 8202 S29 board – inverter control 7623 Topographic diagrams S14 board – line filter 630 S15 board – filament power supply 7961 S17 board – X-ray tube head interface 4617 S19/B board – inverter control 7822 S20 board - X-ray tube stator starter 46 S21/A board – X-ray tube stator rotation 457 S22 board – microprocessor 1401 S25 board – capacitors bank control 1419 S62/A board – keyboard interface 8203 S29 board – inverter control 1420 Exploded views Mechanical 7698 Electrical 7862 Section 6 Page 64 of 65 Doc. 9535 Technical Manual X-RAY SETTINGS Section 5 Doc. 9535 Page 65 of 65