Liquid sense sensitivity adjustment (PLL circuit) SAMPLE ROTOR: • Turn “P1” to the extreme left (CCW) position (very insensitive). • Put a dry and empty Sample tube in the Rotor on position “2”, surrounded by empty paediatric adapters on position’s 1, 3, E3, 4, W, B and S1. • Select in the service menu, sample arm and reset the sample arm [F1]. • Turn the sample arm into the Sample tube on position “2” [→][→][→][↓]. • The liquid detection has to be “NO”. • Turn “P2” to the right (CW) position until the liquid detection changes to a “YES”. Now turn “P2” to the left (CCW) until the bit changes again to a “NO”. • Turn “P2” 360° (one complete turn) to the left (CCW). REAGENT ROTOR: • Turn “P2” to the extreme left (CCW) position (very insensitive). • Put an empty 5ml bottle on position “28”, surrounded by water filled bottles on position’s 15, 24, 25, 26, 27, 29, 30, 31, 32, 1 and 16. • Select in the service menu, reagent arm and reset the reagent arm [F1]. • Turn the reagent arm into the empty bottle on position “28” [←][←][←][←][↓]. The liquid detection has to be “NO”. • Turn “P1” to the right (CW) position until the liquid detection changes to “YES”. Now turn “P1” to the left (CCW) until the bit changes again to a “NO”. • Turn “P1” 360° (one complete turn) to the left (CCW). Check also for some other positions if the liquid sense is working well.