B40/B20 Patient Monitor Technical Reference Manual Patient Monitor B40/B20 English 2081505-001 G (Paper) © 2021 General Electric Company. All Rights Reserved. B40/B20 Patient Monitor Technical Reference Manual 0459 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EC. All specifications subject to change without notice. Order code 2081505-001 Revision G 24 March, 2021 GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI USA Zip: 53223 Tel: 1 414 355 5000 (outside US) 800 558 5102 (US only) Fax: 1 414 355 3790 www.gehealthcare.com Copyright © 2021 General Electric Company. All rights reserved. GE Healthcare 3F Building 1, GE Technology Park 1 Huatuo Road Shanghai PRC 201203 Tel: +86 21 3877 7888 Fax: +86 21 3877 7451 Classifications In accordance with IEC 60601-1 − Class I and internally powered equipment - the type of protection against electric shock. − Type BF or CF equipment. The degree of protection against electric shock is indicated by a symbol on each parameter module. − Equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. − Continuous operation according to the mode of operation. − Portable Monitor In accordance with IEC 60529 − IP21 - degree of protection against harmful ingress of water. In accordance with EU Medical Device Directive − IIb. In accordance with CISPR 11: − Group 1 Class A; • Group 1 contains all ISM (Industrial, scientific and medical) equipment in which there is intentionally generated and/or used conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself. • Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network which supplies buildings used for domestic purposes. Trademarks Listed below are GE Medical Systems Information Technologies and GE Healthcare Finland Oy trademarks used in this document. All other product and company names contained herein are the property of their respective owners. Datex, Ohmeda, DINAMAP, Trim Knob, Unity Network, CARESCAPE, EK-Pro, TruSignal, Entropy, GE Healthcare, GE Medical system, General Electric Company. 1 Introduction About this manual 1 1 4 Overview 1.1 1.2 1.3 1.4 2 System description 2.1 2.2 2.3 2.4 2.5 2.6 3 3.2 3.3 4.3 4.4 16 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 3.1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 3.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 3.1.3 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 3.1.4 Power board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 3.1.5 AC/DC unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 3.1.6 Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Interfacing computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 3.3.1 External connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Hemo-dynamic module introduction 4.1 4.2 13 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Bus structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Distributed processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Module communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.4.1 Serial communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Software loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Frame functional description 3.1 4 Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2.2 Safety message signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2.3 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.2.4 ESD precautionary procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.2.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Service information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.3.1 Service requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.3.2 Equipment identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Product security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 1.4.1 Security features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 1.4.2 Security operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 1.4.3 Product change management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 1.4.4 Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 25 Monitor software compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.1 Hemo-dynamic module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.2 Signals and isolation barrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 4.2.3 Power supply section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 4.2.4 NIBP board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 4.2.5 ECG board in 5-lead measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 4.2.6 STP board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 4.3.1 Front panel connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 i Document no. 2081505-001 B40/B20 Patient Monitor 4.4.1 4.4.2 4.4.3 4.4.4 4.4.5 4.4.6 2 Hardware installation 1 Installation 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 1.10 3 NIBP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Pulse oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 1 Unpacking instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Choosing location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Mounting the monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Connection to power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Check configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Connection to Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.6.1 Pre-installation requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.6.2 To connect the network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.6.3 Network configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Inserting and removing the E modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Monitor connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Visual indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Installation checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Maintenance 1 Instructions 1.1 1.2 2 Electrical Safety Tests 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 3 3 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Power Outlet Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Power cord and plug. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Ground (earth) integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 2.4.1 Ground Continuity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 2.4.2 Impedance of Protective Earth Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Earth leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Enclosure (Touch) leakage current test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Patient leakage current tests - overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 2.7.1 Patient (source) leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.7.2 Patient (sink) leakage current tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Installation checkout 3.1 3.2 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Recommended tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 13 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.2.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 3.2.3 Time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 ii Document no. 2081505-001 3.2.4 3.2.5 3.2.6 3.2.7 4 Maintenance and checkout 4.1 4.2 4.3 5 16 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 4.2.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 4.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.2.3 Keyboard(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.2.4 Time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.2.5 Hemo Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.2.6 Loudspeaker. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.2.7 Monitor software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.2.8 Watchdog circuitry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.2.9 Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.2.10 Final cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Monitor battery maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 4.3.1 Use recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 4.3.2 Storage recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 4.3.3 Test the battery charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 4.3.4 Charge the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 4.3.5 Condition the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 4.3.6 Replace the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Adjustments and calibrations 5.1 5.2 5.3 4 Parameters measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 MC or S/5 Network connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Conclusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 24 NIBP calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Temperature calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Invasive pressure calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Troubleshooting 1 Introduction 1.1 1.2 1.3 2 Frame troubleshooting 2.1 2.2 3 3.5 3.6 3.7 3.8 4 Battery indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 NET section troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Hemo Troubleshooting 3.1 3.2 3.3 3.4 1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 General troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Software troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 9 NIBP troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 ECG troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 STP troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.4.1 NIBP toubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.4.2 NIBP error code explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Impedance respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Pulse oximetry (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 iii Document no. 2081505-001 B40/B20 Patient Monitor 3.9 5 Service Menu 1 Introduction 1.1 1.2 2 3 3.2 3.3 4 5 5.2 5.3 5.4 5.5 5.6 20 Keyboard Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Parameters 5.1 7 Country Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 3.1.1 Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 3.2.1 Network Config (MC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 3.2.2 TCP/IP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 3.2.3 HL7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.2.4 Dri Config (S/5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 3.2.5 Dri Comm (in S/5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 3.3.1 Power page. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 3.3.2 WPM Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Keyboard 4.1 3 SW Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Active Inactive SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Module SW Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Country Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 License. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Enter/Exit Demo Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Frame 3.1 1 Service Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 SW Management 2.1 2.2 2.3 2.4 2.5 2.6 6 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 22 Gas Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 5.1.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 5.1.2 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 ECG Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 5.2.1 ECG Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 STP Module (for GE SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 5.3.1 Calibrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 NIBP Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 5.4.1 NIBP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 5.4.2 NIBP Safety Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 5.4.3 NIBP Pneumatics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 SpO2 (for Masimo/Nellcor SpO2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Entropy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 6 Set/Test 36 7 Service Log 37 Field replaceable unit iv Document no. 2081505-001 1 Spare part 1.1 1.2 1.3 1.4 1.5 1.6 1.7 2 Disassembly 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 2.10 7 1 Front cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Back cover unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Frame. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Extension rack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Hemo box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 FRU parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Other parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Reassembly precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 To separate the frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 To disassemble the frame. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 To disassemble the extended rack and the recorder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Handling and storage of display component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 To replace the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 To download the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 E-miniC Module 1 Product overview 1.1 1.2 1.3 1.4 2 Maintenance and checkout 2.1 2.2 2.3 3 3.2 4 13 Sample flow rate adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.1.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.1.2 Sample flow rate adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Gas calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 3.2.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 3.2.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Troubleshooting 4.1 7 Replacement of planned maintenance parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2.1.1 Required parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2.1.2 Replacement procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Functional checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.3.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.3.3 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Calibration and adjustments 3.1 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2.1 CO2 measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.3.1 Gas sampling system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.3.2 MiniC sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.3.3 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 16 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 v Document no. 2081505-001 B40/B20 Patient Monitor 4.2 4.3 5 Disassembly and reassembly 5.1 5.2 6 19 Disassembly guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 5.1.1 Serviceable parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 5.1.2 Service limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 5.1.3 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 5.1.4 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 5.1.5 Tools needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Disassembly and reassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 5.2.1 Replacing the pump unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 5.2.2 Replacing the miniCO2 assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Service parts 6.1 6.2 8 Troubleshooting checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.2.1 Gas sampling system troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.2.2 MiniC sensor troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Troubleshooting charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 22 Ordering parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Spare parts for E-miniC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 E-sCO, E-sCAiO and N-CAiO Modules 1 Product overview 1.1 1.2 1.3 2 Maintenance and checkout 2.1 2.2 2.3 3 3.2 13 Replacement of planned maintenance parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.1.1 Required parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.1.2 Planned Maintenance Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.1.3 Replacement procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Functional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.3.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.3.3 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Calibration and adjustments 3.1 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 CO2, N2O, and agent measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 O2 measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.3.1 Controls and connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.3.2 Gas sampling system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.3.3 MiniTPX measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.3.4 MiniOM Oxygen sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.3.5 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.3.6 MiniOM board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 1.3.7 Main Component Interactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 21 Sample Flow Rate Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 3.1.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 3.1.2 Sample Flow Rate Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Gas Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 3.2.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 vi Document no. 2081505-001 3.2.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 4 Troubleshooting 4.1 4.2 4.3 4.4 4.5 5 Disassembly and reassembly 5.1 5.2 6 6.2 9 31 Disassembly guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 5.1.1 Serviceable parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 5.1.2 Service limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 5.1.3 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 5.1.4 Protection from dust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 5.1.5 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 5.1.6 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Disassembly and reassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 5.2.1 Disassembly workflow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 5.2.2 Detaching the Front Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.2.3 Detaching the Module Casing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.2.4 Replacement of Planned Maintenance Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.2.5 Replacement of CO2 Absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 5.2.6 Detaching the Latch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 5.2.7 Detaching the Front Chassis Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 5.2.8 Detaching the Main Flow Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 5.2.9 Detaching the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 5.2.10 Detaching the OM holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Service parts 6.1 24 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Troubleshooting checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.1 Gas sampling system troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.2 MiniOM Measuring unit troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.3 MiniTPX Measuring unit troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.4 CPU board troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Service Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 4.4.1 Gas measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Troubleshooting charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 4.5.1 Gas measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 39 Ordering parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 6.1.1 Planned Maintenance Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Spare parts for E-sCAiO, E-sCO, N-CAiO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 6.2.1 Front covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 E-Entropy Module 1 Product overview 1.1 1.2 1.3 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2.1 EntrEEG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 FEMG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.3 RE and SE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.4 Impedance measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.3.1 Controls and connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.3.2 Entropy board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 vii Document no. 2081505-001 B40/B20 Patient Monitor 2 Maintenance and checkout 2.1 2.2 2.3 3 Troubleshooting 3.1 3.2 4 4.2 5 10 Disassembly guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 4.1.1 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 4.1.2 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 4.1.3 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Disassembling and reassembling procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 4.2.1 To replace the front cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 4.2.2 To replace the Entropy board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 4.2.3 Reassembling the module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Service parts 5.1 5.2 8 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Disassembly and reassembly 4.1 5 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Electrical safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Functional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 2.3.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 2.3.2 Entropy tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 2.3.3 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 12 Ordering parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Spare parts for E-ENTROPY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Appendix A: ElectroMagnetic Compatibility A-1 Appendix B: Installation and checkout form, B40/B20 B-1 Appendix C: Maintenance and checkout form, B40/B20 C-1 Appendix D: Service check form, E-miniC D-1 Appendix E: Service check form, E-sCO, E-sCAiO and N-CAiO E-1 Appendix F: Service check form, E-Entropy F-1 Appendix G: Networking disclosure to facilitate network risk managment G-1 viii Document no. 2081505-001 1 Introduction About this manual Indications for use: B40 This device is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The device is intended for use under the direct supervision of a licensed health care practitioner. The device is not intended for use during MRI. The device can be a stand-alone monitor or interfaced to other devices via a network. The device monitors and displays : ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardia l/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate), and Entropy. Indications for use: B20 This device is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The device is intended for use under the direct supervision of a licensed health care practitioner. The device is not intended for use during MRI. The device can be a stand-alone monitor or interfaced to other devices via a network. The device monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardia l/Core/Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion, and Entropy. Intended audience This Technical reference manual is meant for service representatives and technical personnel who install, configure, maintain, administer, troubleshoot or repair B40/B20 monitor running the software license VSP-C. Notes to the reader As the monitor setup may vary, some functions described may not be available in the monitor you are using. • The order code for the manual is 2081505-001. 1-1 Document no. 2081505-001 B40/B20 Patient Monitor • Read the manual through and make sure that you understand the procedures described before the installation of the monitor. To avoid risks concerning safety or health, strictly observe the warning indications. If you need any assistance concerning the installation, please do not hesitate to contact your authorized distributor. Installation without network are allowed by customer. The network installation and service are allowed by authorized service personnel only. GE Healthcare assumes no responsibility for the use or reliability of its software in equipment that is not furnished by GE. Responsibility of the manufacturer GE Medical Systems Information Technologies, Inc. (GE) is responsible for the effects on safety, reliability and performance of the equipment only if: − Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE. − The electrical installation of the relevant room complies with the requirements of the appropriate regulations. − The equipment is used in accordance with the “User's Guide.” − The equipment is installed, maintained and serviced in accordance with this manual. Product availability Some of the product parts and accessories mentioned in this manual may not be available in all countries. Please, consult your local representative for the availability. Related documentation − Clinical aspects, basic methods of measurement and technical background: Patient Monitor B40/B20 User’s Reference Manual − Options and selections of the software: Patient Monitor B40/B20 Default Configuration Worksheet − Compatible supplies and accessories: Patient Monitor B40/B20 Supplies and Accessories − Other devices closely related to the monitor: • • • iCentral and iCentral Client User's Reference Manual CIC Pro Clinical Information Center Operator's Manual CARESCAPE Central Station User’s Manual Manual conventions To help you find and interpret information easily, the manual uses consistent text formats: Sign the check form after performing the procedure. Within this manual, special styles and formats are used to distinguish between terms viewed on screen, a button you must press, or a list of menu commands you must select: − Names of hardware keys on the keypad are written in bold typeface: NIBP Start/Cancel. − Menu items are written in bold italic typeface: Monitor Setup. − Emphasized text is in italic typeface. 1-2 Document no. 2081505-001 − When referring to different sections in this manual, section names are enclosed in double quotes: “Cleaning and care”. − The word “select” means choosing and confirming. − Messages (alarm messages, informative messages) displayed on the screen are written inside single quotes: 'Learning.' − Note statements provide application tips or other useful information. Monitor naming conventions In this manual, the B40 Patient Monitor and B20 Patient Monitor are referred to as “the monitor” when a function or a feature applies to both. For describing monitor-specific issues, the monitors are referred to as B40 and B20 respectively. Acquistion module naming conventions In this manual, the following naming conventions are used to refer to different modules and module categories: • • • • Single-width airway module: E-miniC CARESCAPE respiratory modules: E-sCO, E-sCAiO Airway Gas Option: N-CAiO E-modules: All modules with the prefix E-, In parameter chapters, E-modules refers to those modules that measure the parameter(s) in question. Illustrations and names All illustrations in this manual are only examples, and may not necessarily reflect your system settings or data displayed in your system. If a particular selection is not available in your system, the selection is shown grayed. 1-3 Document no. 2081505-001 B40/B20 Patient Monitor 1 Overview The montior is a modular multiparameter patient monitor. The monitor is especially designed for monitoring in intensive care units. It can also be used during transportation within the hospital. The modular design makes the system flexible and easy to upgrade. NOTE: Your system may not include all these components. Consult your local representative for the available components. 3 1, 4 2 1. The monitor frame 2. E modules: The compatible acquisition modules 3. Extension rack 4. Software: VSP-C 1-4 Document no. 2081505-001 Overview 1.1 Symbols - - On the rear panel: - Electric shock hazard. Do not open the cover or the back. Refer servicing to qualified personnel. - For continued protection against fire hazard, replace the fuse only with one of the same type and rating. - Disconnect from the power supply before servicing. - Do not use without manufacturer approved mounting. On the hemo module: Protection against cardiac defibrillator discharge is due in part to the accessories for pulse oximetry (SpO2), temperature (T) and invasive pressure (P) measurement Follow operating instructions Refer to instruction manual/booklet Electrostatic sensitive device. Connections should not be made to this device unless ESD precautionary procedures are followed. Type BF (IEC 60601-1) protection against electric shock. Isolated (floating) applied part suitable for intentional external and internal application to the patient, excluding direct cardiac application. Type BF (IEC 60601-1) defibrillator-proof protection against electric shock. Isolated (floating) applied part suitable for intentional external and internal application to the patient, excluding direct cardiac application. Type CF (IEC 60601-1) protection against electric shock. Isolated (floating) applied part suitable for intentional external and internal application to the patient, including direct cardiac application. Type CF (IEC 60601-1) defibrillator-proof protection against electric shock. Isolated (floating) applied part suitable for intentional external and internal application to the patient including direct cardiac application. In the front panel: battery Equipotentiality. Monitor can be connected to potential equalization conductor. Alternating current 1-5 Document no. 2081505-001 B40/B20 Patient Monitor Audio pause. Home. Return to the normal screen. ON/OFF. Fuse. Replace the fuse only with one of the same type and rating Gas inlet. Gas outlet. IP21 SN,S/N Degree of ingress protection. Serial number Date of manufacture. This symbol indicates the date of manufacture of this device. The four digits identify the year. Manufacturer. This symbol indicates the name and the address of the manufacturer. European authorized representative. European Union Declaration of Conformity. Rx Only U.S. Prescriptive Device. USA only. For use by or on the order of a Physician or persons licensed by state law. Fragile. Handle with care. Keep dry. Protect from rain. This way up. Storage temperature Humidity limitations. 1-6 Document no. 2081505-001 Overview Atmospheric pressure limitations. Recycled materials or may be recycled. This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please, contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment. The separate collection symbol is affixed to a battery, or its packaging, to advise you that the battery must be recycled or disposed of in accordance with local or country laws. To minimize potential effects on the environment and human health, it is important that all marked batteries that you remove from the product are properly recycled or disposed. For information on how the battery may be safely removed from the device, please consult the service manual or equipment instructions. Information on the potential effects on the environment and human health of the substances used in batteries is available at this url: http://www.gehealthcare.com/euen/weee-recycling/index.html A B B Battery operation and remaining capacity. The height of the green bar indicates the charging level. Battery (A) charging (white bar) Battery (A) failure B No battery backup Battery (A) missing Submenu. Selecting a menu item with this symbol opens a new menu. The monitor is connected to Network. A blinking heart next to the heart rate or pulse rate value indicates the beats detected. A lung next to the respiration rate value indicates that respiration rate is calculated from the impedance respiration measurement. 1-7 Document no. 2081505-001 B40/B20 Patient Monitor Alarm off indicator - Displays in the digit field and in the Alarms Setup menu when physiological alarms for this parameter are turned off. Audio alarms off indicator - Displays in the upper left corner of the alarm area when physiological audible alarms are turned off. Audio alarms paused indicator - Indicates all audio alarms are paused and the amount of time remaining for the alarm pause period displays as a countdown timer. Displays in the upper left corner of the screen. 1.2 Safety information 1.2.1 General This device is intended for use under the direct supervision of a licensed health care practitioner. Contact GE for information before connecting any devices to the equipment that are not recommended in this manual. Parts and accessories used must meet the requirements of the applicable IEC 60601 series safety standards, and/or the system configuration must meet the requirements of the IEC 60601-1 medical electrical systems standard. Periodically, and whenever the integrity of the device is in doubt, test all functions. The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include: • • use of the accessory in the PATIENT VICINITY; and evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601-1. If the installation of the equipment, in the USA, will use 240V rather than 120V, the source must be a center-tapped, 240V, single-phase circuit. 1.2.2 Safety message signal words Safety message signal words designate the severity of a potential hazard. DANGER: Indicates a hazardous situation that, if not avoided, will result in death orserious injury. No danger messages apply to this system. WARNING: Indicates a hazardous situation that, if not avoided, could result in death orserious injury. CAUTION: Indicates a hazardous situation that, if not avoided, could result in minor or moderate injury. NOTE: Indicates a hazardous situation not related to personal injury that, if not avoided, could result in property damage. 1-8 Document no. 2081505-001 Overview 1.2.3 Safety precautions The following list contains general warnings and cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User’s Guide and User’s Reference Manual. Warnings • Use only GE recommended power cords. • When disconnecting the system from the power line, remove the plug from the wall outlet first. • • • • • • • • • • • • • • • • • • • • • Due to high voltage, use insulated screw driver. High voltage on test body; do not touch it during the test. To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth. Always check that power cord and plug are intact and undamaged. All system devices must be connected to the same power supply circuit. Only interconnect devices when determined safe by qualified biomedical personnel. Only devices that are specified compliant with IEC 60950-1 or IEC 60601-1 may be connected to the Ethernet MC ports. Biomed must determine interconnected parts are safe. Do not use without manufacturer approved mounting. If the software package is changed, all clinical settings will reset to factory defaults. Do not use with iCentral software V5.0.3 and earlier. Do not use Mobile Care Server software V5.2 and earlier. Verify compatibility of all system components prior to installation. Use only approved accessories, including mounts, and defibrillator-proof cables and invasive pressure transducers. For a list of approved accessories, see the supplies and accessories list delivered with the monitor. Other cables, transducers and accessories may cause a safety hazard, damage the equipment or system, result in increased emissions or decreased immunity of the equipment or system or interfere with the measurement. Regular preventive maintenance should be carried out every 24 months. Do not use multiple modules with identical measurements in the same monitor. The user may only perform maintenance procedures specifically described in this manual. Incorrect power line frequency setting could adversely affect ECG processing. Make sure patient is not being monitored while servicing the equipment. Don’t press power key when changing language or doing factory reset. Never store the monitor with the batteries inside. Storing the monitor with the batteries inside may result in damage to the monitor. 1-9 Document no. 2081505-001 B40/B20 Patient Monitor Cautions • Set the time of a newly added network device as close as possible to the time of devices already on the network. 1.2.4 ESD precautionary procedures • To avoid electrostatic charges building up, it is recommended to store, maintain and use the equipment at a relative humidity of 30% or greater. Floors should be covered by ESD dissipative carpets or similar. Non-synthetic clothing should be used when working with the component. • To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the equipment, one should touch the metallic frame of the component or a large metal object located close to the equipment. When working with the equipment and specifically when the ESD sensitive parts of the equipment may be touched, a grounded wrist strap intended for use with ESD sensitive equipment should be worn. Refer to the documentation provided with the wrist straps for details of proper use. ESD precautionary procedure training It is recommended that all potential users receive an explanation of the ESD warning symbol and training in ESD precautionary procedures. The minimum contents of an ESD precautionary procedure training should include an introduction to the physics of electrostatic charge, the voltage levels that can occur in normal practice and the damage that can be done to electronic components if they are touched by an operator who is electrostatically charged. Further, an explanation should be given of methods to prevent build-up of electrostatic charge and how and why to discharge one’s body to earth or to the frame of the equipment or bond oneself by means of a wrist strap to the equipment or the earth prior to making a connection. 1.2.5 Disposal Dispose of the whole device, parts of it and its packing material and manuals in accordance with local environmental and waste disposal regulations. 1.3 Service information 1.3.1 Service requirements Follow the service requirements listed below. • • • Refer equipment servicing to GE authorized service personnel only. • Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards. • Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required. Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user's responsibility to report the need for service to GE or to one of their authorized agents. 1-10 Document no. 2081505-001 Overview 1.3.2 Equipment identification Every GE device has a unique serial number for identification. The device plate is located on the rear of the patient monitor. Serial number for B40: SKZYYWW****WA Serial number for B20: SM2YYWW****WA 1.4 Product security The patient monitoring software incorporates an assortment of security features designed to allow a flexible approach to safe and secure implementation, focusing on the principles of confidentiality, integrity, and availability. These features assist you in using the system in a manner that protects patient privacy and security in your setting, and also addresses expectations for the environment where the system will be used. 1.4.1 Security features Access control Access control is the overall mechanism used to determine and enforce the following: • • • • Who has access How individuals gain access When access is permitted What information may be accessed Other than clinical and service applications, access to other subsystems (for example bootloader) is restricted. The clinical and service application interfaces have a role-based access control (for example, biomed and clinical). A user may log into these interfaces to perform operations that are limited to the generic user. See the user and technical manuals for detailed information on available features. Authentication Authentication is the process of proving individual identity, and is a key element in an access control system. In the clinical and service applications, there are certain features that require user authentication. To access these features, the user must log into the clinical and service applications with a valid username and password. Authorization Authorization is the process of granting and revoking access to information, and is another key element in an access control system. Although primarily an administrative process that is driven by an organization’s policies and procedures, the patient monitor contains features that will help implement and enforce an organization’s method. The applications have an authorization mechanism to provide information to the user. Audit The ability to record and examine system activity is crucial to a successful information security program, as well as a regulatory requirement in most environments. The patient monitor stores system logs. 1-11 Document no. 2081505-001 B40/B20 Patient Monitor Malicious software protection Vigilant defense on many levels is required to keep systems free from compromise by malicious software. Effective protection requires cooperation and partnership between GE and our customers. The following product features contribute to defense against malicious software: • Device design and configuration (hardening) The patient monitor has been hardened through the restriction and removal of user access to core operating system functionality. In addition, unneeded functionality has been removed or restricted. • Antivirus software To provide seamless real-time patient monitoring, the patient monitor does not have antivirus software. • Security updates and patching processes Security updates and patches cannot be applied to the product without going through GE’s vigorous software verification and validation process. Any software update needs will be communicated by GE. 1.4.2 Security operations Network security GE requires that the MC port of the patient monitor be connected to a physically or virtually dedicated CARESCAPE Network MC, S/5 Network or HL7 Network, isolated from all other networks. The monitor uses a defense-in-depth approach protecting the monitor from unauthorized access via the network. This layered approach leverages internal software firewall and network isolation. 1.4.3 Product change management GE has rigorous software verification and validation processes. Any software update needs will be communicated by GE. The patient monitoring system, including all aspects of software, should be used as it was intended by GE. 1.4.4 Communication For detailed product security information, go to one of the following Web addresses: http://www.gehealthcare.com/usen/security http://www.gehealthcare.com/usen/security/mds2.html 1-12 Document no. 2081505-001 System description 2 System description 2.1 Introduction The monitor build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed. 2.2 Bus structure The operation of monitor is based on two communication channels, the CPU bus and module bus. All units, including the modules, receive power from the same power supply, which is an integral part of the monitor frame. NAND Flash Ethernet DATA BUS SDRAM AT91 ARM USB HOSE AT91SA M7s256 IIC BUS Sound RS485 Address bus LCD Figure 1 LCD DATA BUS General bus structure of monitor The CPU bus is a communication channel used only for internal data transfer. It is based on the AT91 ARM local bus. Data and address are transferred on this 32 bit wide bus using the CPU clock frequency. The module bus is for the parameter modules. The bus is based on the industry standard RS-485, which uses a differential serial method to transfer data. The module bus uses a 500 kbps data transfer rate. The RS-485 type of serial communication supports so-called multidrop or party line connections. This means that all parameter modules connected to the module bus use exactly the same lines for communication. The advantage of this is that all bus connectors are identical and the modules can be connected in any order and position. 1-13 Document no. 2081505-001 B40/B20 Patient Monitor 2.3 Distributed processing This is a multiprocessor system. All parameter modules have their own microprocessor, which performs functions such as waveform filtering, parameter related computing and pneumatic control, etc. At the same time the main CPU performs higher level tasks such as trending and alarm control. While the parameter modules and CPU are performing their tasks, the UPI (Universal Peripheral Interface) microprocessor handles all functions needed to transfer data between the parameter modules and the CPU. This kind of parallel processing gives one major advantage to centralized processing. When new parameter modules are added to the system, the processing power is increased. As a result, the system does not slow down when new features are added. 2.4 Module communication The communication master controlling data transfers between the CPU and parameter modules is called UPI processor. It sends data to each connected module 100 times a second. Modules respond to each data request immediately by sending a data package, whose length depends on the type of the module. This communication protocol ensures that each module receives and sends data every 10 ms. If a module does not respond to data requests, the UPI processor presumes that the module is disconnected. The data transfered on USB bus between main CPU and UPI processor. Marker Out Main CPU Figure 2 USB BUS UPI Processor Module BUS Principle of UPI section operation 2.4.1 Serial communication An RS485 type bus driver makes the serial communication between the module and the frame. The data transmission rate is 500kbps. Module Bus (RS-485) Frame Hemo module Figure 3 1-14 Document no. 2081505-001 Recorder E-module Serial communication between Frame and modules System description 2.5 Parameter modules PATIENT A/D convert Peripheral drivers Figure 4 CPU +13...16V VMOD RAM EEPROM Opto isolation Data MODULE BUS Analog electronics Isolation transformer Patient isolation +5V RS485 drivers General structure of parameter modules with patient isolation The detailed structure of a parameter module depends on the specific needs for each individual parameter. However, some common parts are used in the parameter modules. The electronics inside the module is usually divided into isolated (floating) and non-isolated sections. Typically, the non-isolated section consists of buffers to interface the parameter module to the module bus while the rest of the electronics is located in the isolated section. The isolated section includes the microcontroller together with memory components, the front-end analog electronics (amplifiers, etc.) and sensor drivers. 2.6 Software loading The program memory on the CPU board is loaded with monitor software and selected language files at the factory. The software is used for running all the functions that are integrated into the CPU board. For service upgrade main software and language files, please refer to the 2062416-001: B40/B20 Patient Monitor Software download instruction. How to do cold start? The patient monitor performs a cold start, if there is over 15 minutes from the previous power off. You can perform a cold start by 2 methods: • Press ON/OFF button to turn off the monitor, waiting for 15 minutes to turn on the monitor. Or, • Press ON/OFF button for about 15 seconds (until the words “monitor is shutting downing...“ disappear) to turn off the monitor. Then turn on the monitor. NOTE: All the patient data and monitor settings will be lost after cold start. 1-15 Document no. 2081505-001 B40/B20 Patient Monitor 3 Frame functional description 3.1 Main components 3.1.1 Keypad User interface parts The Horizontal Membrane keypad containing 20 keys. The keypads are foil membrane keypads. The keypads are connected to the UPI section of the CPU board. Trim Knob is used for menu selection. 3.1.2 Display The B40 use 12.1” LED backlight display with SVGA 800 x 600 resolution has bright long life light bar and a wide viewing angle. The B20 use 10.4” LED backlight display with SVGA 800 x 600 resolution has bright long life light bar and a wide viewing angle. Backlights The backlight of LCD is made up of 1 light bar, LED to be 3020, 20 pieces, 10 serials and 2 parallels. The backlight unit is driven by a separate driver board. 1-16 Document no. 2081505-001 Frame functional description I/O connector AC INLET 100-240 Vac 50/60 Hz Cable or wires Pin-to-pin connection Alarm light board Speaker AC/DC Unit LCD display SVGA Nurse call Power Board LED driver board CPU Board Serial data XY/CW Module bus Module bus Module interface board Network Vmod BAT2 SMBUS BAT1 SMBUS User interface board External interface board Network Smart battery B Li-ion 9-12.6V Module bus connector CW Battery board Smart battery A Li-ion 9-12.6V Module bus connector XY Trim Knob Keyboard/Memb rane switch Power indicator Multi I/O Multi I/O adaptor Nurse DFB Serial call MK out port Figure 5 Block diagram 1-17 Document no. 2081505-001 B40/B20 Patient Monitor 3.1.3 CPU board The board is based on AT91 ARM microprocessor. Other functions include LVDS display driver, 10/100Mbps on board Ethernet, WLAN communication, Alarm Light function, KEY board and rotor encoder control, audio driver function, nurse call function, defibrillation function, module bus function. The CPU section takes care of the central processing. The main features are: • • • • • • AT91 ARM 266 MHz Main CPU clock 64MBytes SDRAM 128 MBytes minimum NAND flash memory 4 MBytes Data flash memory Main CPU Provides one standard UART communication Connectors Ethernet communication connector WLAN card operation connector Color LED baclight display operation connector Audio operation connector Alarm Light indicator operation connector Power board connection connector Voltage supervision There are two voltage supervision chips that control the system reset signals. The +3.3V supervision chip outputs reset signals for +3.3V devices. 3.3V Reset Threshold will be Falling: min 3.00 V; max 3.15 V. 1-18 Document no. 2081505-001 Frame functional description The +1.2V supervision chip outputs reset signals for +1.2V devices. 1.2V Reset Threshold will be Falling: min 1.08V; max 1.14 V. 3.1.4 Power board The Power board converts the output voltage of AC/DC unit and battery voltage to various supply voltages for the electronics of monitor. The Power board provides monitor system power function, module bus power function, display backlight power and power failure alarm. Power board operation is controlled by PMC (Power Management Controller) CPU. PMC takes care of power path controlling. Power Board incorporates the SMBus interface between the PMC, battery charger IC, and smart battery. Power board provides the system voltage for +3.3V, +5V. And Power Board provides 15V power for measurement modules connected patient monitor module bus. The Boost converter of PMC system provides module voltage supply for measurement modules through patient monitor module bus.It operates at input voltage from 9V to15V. Power Board create an power failure alarm for user to notice unexpected loss of power supply. It indicate power failure alarm by blinking patient monitor yellow alarm light and buzzer alarm. Block diagram of the power supplies is represented in following 1-19 Document no. 2081505-001 B40/B20 Patient Monitor MUX +5V_PMC ACDC (15V) Battery Charger Battery 1 Power Path Power Management Controller (PMC) VSYS Enable Switch Over Voltage Protection (Crowbar) Battery 2 VSYS +3.3V 3.3V & 5V Buck Regulator 15V Boost Regulator Figure 6 Power Fail Alarm +5V Over Current Protection & Current Sense VMOD +5V_PMC Linear Regulator +5V_PMC LCD Backlight Enable Switch Backlight_VCC Power board block diagram Power management controller The power management controller (PMC) used is the Freescale MC9508AC60CFUE, 8 bit microcontroller. The PMC is used to: − Control power supplies sequencing − Monitor the power supply voltages and currents via internal analog to digital converters (ADCs) − Disabling power supplies during fault conditions − Communicate with the CPU board via UART communication − Read and write to a IIC EEPROM − Communicate with smart batteries and a level 2 smart battery charger via SMBus. 1-20 Document no. 2081505-001 Frame functional description 3.1.5 AC/DC unit AC/DC unit The AC/DC unit is a compact medical power supply based on high-efficiency technology. It is designed for 65 watt continuous output power, universal AC input and 15V output voltage. AC/DC insulation diagram as follow: 3.1.6 Batteries The monitor has two lithium-ion batteries, located in the battery compartment. The power board connects one of the batteries to be the power source, if no power is received from the AC/DC unit.The battery charging is controlled by the power board. The batteries can be charged separately, and screen symbols and monitor frame LED indicators indicate their charging level and possible failure. NOTE: When the monitor is battery powered, the green battery LED is on. When the monitor is mains powered, the green mains LED is on. 3.2 Interfacing computer A computer is connected to the serial port connector on the Multi I/O adapter. Contact your authorized GE Healthcare distributor for further advice on computer interface. WARNING Connecting electrical equipment together or using the same extension cord for more than one device may cause their leakage currents to exceed the limits specified in relevant safety standards. Always make sure that the combination complies with the international safety standard IEC 60601-1 for medical electrical systems and with the requirements of local authorities. 1-21 Document no. 2081505-001 B40/B20 Patient Monitor WARNING Connecting the power supply cord of the computer to the wall power outlet may cause the computer leakage current to exceed the limit specified for medical equipment. A computer must be supplied from an additional transformer providing at least basic isolation (isolating or separating transformer). 3.3 Connectors and signals 3.3.1 External connectors 1 4 6 3 Figure 7 5 7 2 External connectors of Frame (1) Receptacle for power cord (2) Serial port (3) Defibrillator connector (4) Nurse call connector (5) Network connector (6) Equipotential connector (7) Multi I/O connector NOTE: 2,3,4 is on the multi I/O adpter. Network connector RJ45 connector 1 2 3 4 5 6 7 8 1-22 Document no. 2081505-001 Pin Signal 1 2 3 4 5 6 7 8 Tx + Tx Rx + N/C N/C Rx N/C N/C Frame functional description Multi I/O connector(26 pin) 26 pin female connector 18 26 10 19 Pin Signal 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 GND NC NC NC NC NC SERIAL_TXD SERIAL _CTS# GND GND NC NC DEFIB_MARKER_OUT NC NC 26 GND Pin Signal 1 2 3 GND NC Nurse_Call SERIAL_RXD SERIAL_RTS# GND GND NURSE_CALL NC NC NC NC SERIAL_+3V3 Nurse Call (pin 3) Nurse call connector Recommended cable design: Pin 1 Pin 2 Pin3 1-23 Document no. 2081505-001 B40/B20 Patient Monitor Serial port 9 pin female connector Pin Signal 1 2 3 4 5 6 7 8 9 GND SERIAL_TXD SERIAL_RXD N/C GND N/C SERIAL_CTS SERIAL_RTS N/C Pin Signal L Live PE Protected earth N Neutral Female mini din7 connector Pin Signal 5 1 9 6 Main power Mains connector Defib connector (Pin 7) 1 2 3 4 5 6 7 1-24 Document no. 2081505-001 GND GND GND GND DEFIB_MARKER_OUT NC GND Hemo-dynamic module introduction 4 Hemo-dynamic module introduction The hemo module provide general hemodynamic parameters. Figure 8 Hemo module 4.1 Monitor software compatibility − The monitor using software VSP-C 4.2 Main components 4.2.1 Hemo-dynamic module 1 2 3 1. InvBP connector 2. Temperature connector 3. SpO2 connector 4. ECG connector 5. NIBP connector 4 5 Figure 9 Front panel of hemo module The Hemo-dynamic module including the NIBP measurement, 5-lead ECG with the Impedance Respiration measurement, SpO2 with the plethysmographic waveform, two invasive pressure measurements (IBP1 and IBP2) and two temperature measurements (T1 and T2). 1-25 Document no. 2081505-001 B40/B20 Patient Monitor The monitor displays waveforms and measurement readings, and handles the trending and alarm management. The ECG (e.g. heart beat and arrhythmia detection) and the Impedance Respiration algorithms are in the monitor software. The modules measure signals and send them to the monitor. The NIBP, SpO2, Temperature and Invasive Pressure algorithms are in the module. There are four parameter circuit boards inside the hemo-dynamic module for processing the measurement signals. Each processing board has a microcontroller with software. The NIBP parameter measurement requires one signal processing board, pneumatic system, valve and pump unit connected to NIBP parameter board. The second parameter board is the optional board, for Nellcor or Masimo SpO2 measurement, it’s Masimo MS-2011 board or Covidien NELL1GE-S board at different configuration. The third parameter board is for GE SpO2, IBP and Temperature measurement including input board. All these three parameter is optional, according to different configuration, using different board: it’s STP board, TP board for Nellcor, TP board for Masimo, GE SpO2 board. The fourth parameter board is for 3/5-lead ECG with the Impedance Respiration measurement including ECG input unit connected to the ECG parameter board. All parameter boards are connected together via module bus flex board connecting voltage and module communication, the module communicates with frame through RS-485 bus. 4.2.2 Signals and isolation barrier The communication signals transfer over the isolation barrier by using high isolation voltage (6kV) opto isolators. 4.2.3 Power supply section The power for the electronics on the floating part of the STP and the ECG boards is made on each board with the switching power supplies connected to a high voltage isolated transformer. The switching power supplies on the STP and ECG boards are synchronized to the frequency, about 172.8 kHz of the switching power supply on the NIBP board. The NIBP board supplies non-isolated 5 V to the ECG and STP boards. The module uses only Vmod 15V +/- 3% of the frame. The other voltages of the measuring boards are made by the switching power supplies and regulators or the linear regulators. Each measuring board is protected against overloading with PTC type automatic fuses. 1-26 Document no. 2081505-001 Hemo-dynamic module introduction 4.2.4 NIBP board EEPROM 1024Bytes Module bus connector Main CPU AT91SAM7S256 RS485 interface Pneumatic control Pump connector Valves connector 256KBytes Flash 64KBytes SRAM 10bits ADC PWR_ SYN Pressure sensor NIBP_+5V 6VD Power supply MAIN_REF Safety CPU MSP430F2013 2KB+256B Flash 128B RAM 16bits SigmaDelta ADC Figure 10 NIBP board functional block diagram Signal processing Two signals from the pressure transducers are amplified and sent to the A/D converter. After the converter, digitized signals are sent to the microprocessor for data processing. The NIBP board is controlled with an ARM7 microprocessor at 16 MHz oscillator frequency. Memory The NIBP program memory (processor flash memory) size is 256k x 8. The processor has 64 kBytes RAM. The EEPROM size is 8K x 8 and it is used to store the calibration values for the pressure transducers, the pulse valve constants gained during measurements, the PC board identification, and the module serial number. Software control The software controls valves and a pump. In addition to the individual on/off signals for each component there is a common power switch for the valves and the pump that can be used at pump/valve failures. Safety circuit The NIBP board is equipped with an independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than the preset maximum measurement time, or if the pressure of the cuff is inflated over the specified pressure limit. The maximum measurement time values and pressure limits for different measurement modes have been specified in the technical specification section of this manual. 1-27 Document no. 2081505-001 B40/B20 Patient Monitor Pneumatics 1 7 5, 6 3 2 4 1 The module has the following pneumatics parts: 5. Intake air filter; for preventing dust and other parts from entering the air pump and the valves. 6. Air pump; for pumping the measuring pressure of the cuff. 7. Deflation Valve; for producing a linear pressure fall (bleeding) in order to measure the blood pressure of the patient. 8. Safety valve/Dump valve; The Safety valve/Dump valve is intended to be used for deflating the cuff in single fault case, i.e. to prevent too long a measurement time or too high an inflation pressure of the cuff. 9. Main pressure sensor; for measuring the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial wall movement. 10. Second pressure sensor; for detecting the, cuff loose, cuff occlusion situations, etc. and for recognizing the pressure sensor fault. 11. Cuff connector; for connection and hose identification. Power supply section of the NIBP board All connections are established via a 10-pin connector (male). The module needs a +15 V (dirty) power supply to operate. The supply voltage Vmod 15V +/- 3% is generated in the power supply section of the monitor. The other voltages needed for the operation of the NIBP measurement are made on the NIBP board. The NIBP power supply synchronizes the ECG and STP isolation power and supplies non-isolated 5 V to the ECG and STP board. 1-28 Document no. 2081505-001 Hemo-dynamic module introduction 4.2.5 ECG board in 5-lead measurement The ECG measurement consists of the functions shown in Figure 11. All functions are located in the ECG board except the ECG input unit. PATIENT AND ECG ELECTRODES ECG CABLE - ECG LEAD SET - ECG TRUNK CABLE ECG INPUT UNIT - ECG CONNECTOR - INPUT PROTECTION RESISTORS INPUT PROTECTION DIODES FOR ECG & RESPIRATION MEASUREMENT INPUT FILTERING FOR ECG & RESPIRATION MEASUREMENTS RESPIRATION MEASUREMENT AMPLIFIERS RESPIRATION MEASUREMENT CURRENT SUPPLY ECG PREAMPLIFIERS & RLD CIRCUIT LEADS OFF & PACER & DEFIBRILLATION DETECTION BASELINE RESTORATION RS 485 COMMUNICATION POWER SUPPLY ISOLATION ISOLATION MODULE BUS CONNECTOR Figure 11 NV MEMORY ECG BLOCK DIAGRAM 12_lead_ECG_meas_blck_dgrm.vsd ECG CPU ECG measurement block diagram ECG input unit The ECG input unit consists of the front panel connector and the ECG input connector board with the high voltage protection resistors. The connector for the ECG cable is a green 11-pin rectangle shaped connector. Input protection and filtering The input protection is implemented with high voltage protection resistors in the ECG input unit and with protection diodes in the ECG board. The input filtering for ECG measurement is done with passive RC filtering. ECG preamplifiers The buffer amplifiers are used for each lead. The “Leads off” detection is implemented by measuring the output level of the input buffer amplifiers with the A/D converter of the CPU. The ECG signals are measured using differential amplifiers. 1-29 Document no. 2081505-001 B40/B20 Patient Monitor ECG amplifiers and baseline restoration The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore the baseline of the signal in the middle of the display after the change of the signal level, e.g. after the change of the DC offset voltage. Pacemaker detection Pacer detection has been made by using four slew rate detector circuits. The pacer detection amplifiers have been realized at the front of the slew rate detectors independently of the ECG measuring channels. Respiration impedance supply The 31.25 kHz sine wave generator is used as the respiration measurement signal supply. Analog switches are used for connecting the sine wave to the ECG leads to be measured. Respiration impedance amplifiers Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting the measurement leads. There are also additional amplifiers for increasing the respiration signal gain. When ECG measurement is 5-lead, the respiration measurement is always done between R and F, independently on the ECG lead selection. When ECG measurement is 3-lead, then the respiration measurement is happened at the same lead as the ECG measurement (I, II or III). ECG CPU The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz. RS485 communication The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver circuits are optically isolated from the processor of the module. Power supply The ECG board has a driver-controlled half-bridge switching power supply with 5 kV isolation. The supply voltages have been regulated with linear regulators. ECG filtering B40 monitors have three ECG filtering modes: MONITORING DIAGNOSTIC ST FILTER 0.5 to 40 Hz 0.05 to 150 Hz 0.05 to 40 Hz The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory) movement artifacts. The monitor filter is used in normal monitoring. The diagnostic filter is used if more accurate diagnostic information is needed. The ST filter gives more accurate information of ST segment, but reduces high frequency noise. The high-pass filters 0.5 Hz and 0.05 Hz are done with software. The monitor sends a command to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is to be used. The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz correspondingly. They are software based filters used for the mains supply filtering. In diagnostic mode the upper frequency is 150 Hz and it is limited by software. 1-30 Document no. 2081505-001 Hemo-dynamic module introduction 4.2.6 STP board STP module measures SpO2, two channels of temperatures and two channels of invasive blood pressures. The SpO2 measurement is made optically with an infrared light, a red light sources, and a photosensitive detector. The SpO2 value and Pulse Rate are calculated based on the signals, which are measured with the photosensitive detector in the SpO2 sensor. There are three configurations of SpO2: GE SpO2, Masimo SpO2 and Nellcor SpO2. There are four kinds of STP parameter board: − GE SpO2 board − STP board, integrated GE SpO2, Temperature and pressure − TP board, integrated Temperature and pressure, communicated with Masimo OEM SpO2 through UART port to get SpO2 data and send to host. − TP board, integrated Temperature and pressure, communicated with Nellcor OEM SpO2 through UART port to get SpO2 data and send to host. The temperature measurement is designed for use with YSI-400 series NTC sensors. The Invasive Pressure measurement is designed for use with the bridge type medical pressure sensors. Refer to the following block diagram. PATIENT AND SpO2 PROBE PATIENT AND TEMPERATURE SENSOR PATIENT AND INVASIVE CANNULA OR CATHETER SpO2 TRUNK CABLE TEMPERATURE CONNECTOR FLUSHING KIT & INVASIVE PRESSURE SENSOR WITH DOME SpO2 CONNECTOR INPUT PROTECTION CIRCUITRY INV.PRESSURE CONNECTOR SpO2 PROBE RECOGNITION & LED DRIVE SELECTION MATRIX TEMPERATURE MEASUREMENT AMPLIFIER SpO2 LED DRIVE SENSOR SIGNAL CURRENT SOURCE INV: PRESSURE MEASURE MENT AMPLIFIER SENSOR SIGNAL VOLTAGE SOURCE SpO2 AMPLIFIER A/D CONVERSION RS 485 COMMUNICATION POWER SUPPLY ISOLATION ISOLATION MODULE BUS CONNECTOR Figure 12 NV MEMORY STP BLOCK DIAGRAM STP_brd_blck_dgrm.vsd STP CPU STP block diagram 1-31 Document no. 2081505-001 B40/B20 Patient Monitor PATIENT AND SPO2 PROBE PATIENT AND TEMPERATURE SENSOR PATIENT AND INVASIVE CANNULA OR CATHETER SPO2 TRUNK CABLE TEMPERATURE CONNECTOR FLUSHING KIT & INVASIVE PRESSURE SENSOR WITH DOME SPO2 CONNECTOR INPUT PROTECTION CIRCUITRY INV.PRESSURE CONNECTOR TEMPERATURE MEASUREMENT AMPLIFIER SENSOR SIGNAL CURRENT SOURCE INV: PRESSURE MEASUREMENT AMPLIFIER SENSOR SIGNAL VOLTAGE SOURCE MASIMO OR NELLCOR SPO2 MODULE A/D CONVERSION TEMPERATURE AND IBP CPU NV MEMORY POWER SUPPLY ISOLATION RS 485 COMMUNICATION ISOLATION TP +Masimo or Nellcor SPO2 BLOCK DIAGRAM MODULE BUS CONNECTOR Figure 13 TP board block diagram Microprocessor unit The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz. High speed I/O is used to obtain a pulse control sequence necessary for pulse oximetry measurement. Timing for the clock is from the oscillator. 1-32 Document no. 2081505-001 Hemo-dynamic module introduction Temperature measurement unit The NTC-resistor value in the probe depends on the patient’s temperature. It is measured with the following principle described below. The constant current source is supplied about 38 μA current through the temperature sensor (YSI 400-series NTC resistor). The constant current is caused a voltage over the temperature sensor (NTC resistor). The voltage over the temperature sensor is amplified in a differential amplifier stage. The amplified voltage is transferred to a controller of the STP board through an A/D converter. constant current source Defibrillation/ESD protection resistors and diodes Temperature sensors T1, T2 Ref1 Ref2 T2 R R R 0C:7k36 15C:3k54 25C:2k53 38C:1k30 45C: 984 R R Figure 14 Differential amplifier d/dt 0 0 0 To A/D converter 0 PSM_temp_meas_principle.vsd T1 R Temperature measurement principle Invasive blood pressure measurement unit An isolated +5 V voltage is supplied to the pressure transducer. The differential voltage, which depends on the pressure and the supplied voltage, is calculated from the bridge connection (see the formula below). Uout = Uin × pressure × 5 μV, where Uin is 5 V Uout = 25 μV × pressure [mmHg] Pressure amplification is realized in the instrumentation amplifier. The gain of the amplifier is set to keep the level of the signal transferred to the A/D converter within the measurement range even when there are circumstantial offsets or offsets caused by the transducer. There is a filter before the amplifier to attenuate high frequency disturbances. 1-33 Document no. 2081505-001 B40/B20 Patient Monitor Vin Current measurement G to AD converter Pressure transducer Vout Input filter G to A D converter Figure 15 PSM_pressure_meas_principle.vsd Instrum entation amplifier Pressure measurement principle Pulse oximetry measurement section LED control signals The D/A converters of the microcontroller on the STP board set the LED intensity adjustment values for the infrared and red LEDs of the SpO2 probe. The microcontroller on the STP board switches ON (to the adjusted intensity) and OFF the SpO2 probe LEDs according to the predetermined sequence. LED driving circuit Differential amplifiers measure the LED currents (LED current indication) of the SpO2 probe over the shunt resistors placed in the LED current paths. The LED driving voltages (LED voltage indication) are measured from the driver circuitry. The LED driving circuits also have MOSFET transistor matrix to enable the use of different probe configurations. Measured signal preamplification The preamplifier is a bipolar/single-ended current-to-voltage converter with adjustable gain. A higher gain is used for measuring thin tissue. The preamplification stage has also ambient light reduction and a second amplifier stage. 1-34 Document no. 2081505-001 Hemo-dynamic module introduction LED Driving circuit 1 LED Intensity adjustment 1 LED ON/OFF control 1 LED voltage indication 1 LED current indication 1 LED Driving circuit 2 LED Intensity adjustment 2 LED ON/OFF control 2 LED voltage indication 2 SpO2 Probe LED current indication 2 LP Oximeter channel 1 Gain=7.5 DCsuppression LP Oximeter channel 2 Gain=7.5 Ambient reduction LP Amplifier: Gain = 2 Oximeter channel 3 Gain=7.5 DCsuppression Oximeter channel 4 LP DE-MUX Analog Figure 16 Amplifiers Digital Spo2_measurement_blck_diagr.vsd Preamplifier: Current-to-voltage type Bipolar/single-ended modes Adjustable gain DCsuppression GE Pulse oximetry measurement block diagram Red and infrared channel separation It is possible to multiplex the detector signal to four different channels depending on the content of the signal. The detector signal must at least multiplex into infrared and red signals. Other channels are e.g. for diagnostic purposes. 1-35 Document no. 2081505-001 B40/B20 Patient Monitor 4.3 Connectors and signals 4.3.1 Front panel connectors Table 6 ECG connector ECG Connector Table 7 Document no. 2081505-001 Signal Name 1 2 3 4 5 6 7 8 9 10 11 R/RA; Right arm electrode C2/V2; Chest electrode C3/V3; Chest electrode L/LA; Left arm electrode N/RL; Neutral/Right Leg Drive electrode C1/V1; Chest electrode C4/V4; Chest electrode F/LL; Left Leg electrode C6/V6; Chest electrode C5/V5; Chest electrode Cable Shield SpO2 connector SpO2 connector 1-36 Pin No. Pin No. Signal Description 1 DET_A Photodiode anode 2 DET_C Photodiode cathode 3 DATA- 4 Wire 1/3 LED connection 5 IR_C IR LED cathode 6 OUTER SHIELD 7 DET_SHIELD 8 PRB_ID Bin/ID Resistor+ 9 Wire 3/5 LED Connection 10 RED_C RED LED cathode 11 DATA+ Hemo-dynamic module introduction Table 8 Invasive blood pressure connectors (IBP1, IBP2) Invasive blood pressure connectors (Dual BP) Table 9 Pin No. Signal Description 1 BP_+VREF BP transducer excitation voltage, channel 1 2 BP SIG+ 3 BP_+VREF BP transducer excitation voltage, channel 2 4 AGND Analog ground 5 BP SIG+ BP transducer signal positive (+), channel 2 6 SHIELD BP cable shield 7 AGND Analog ground 8 BP SIG1 BP transducer signal negative (-), channel 1 9 BP SIG2 BP transducer signal negative (-), channel 2 10 BP1_ID BP1 probe identification 11 NC Not connected BP transducer signal positive (+), channel 1 Temp connector (T1, T2) Temp connector Pin No Signal 1 Sensor drive current 2 Input from temperature sensor, channel 1 3 Not connected 4 Not connected 5 Thermistor ID (LOW= Temperature error, HIGH=YSI 400 series) 6 Cable shield 7 Analog ground 8 Input from temperature sensor, channel 2 9 Not connected 10 Not connected 11 Digital ground 1-37 Document no. 2081505-001 B40/B20 Patient Monitor 4.4 Measurement principle 4.4.1 NIBP NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure. The NIBP parameter conforms to EN1060-1:1995/A2:2009 Specification for Non-invasive sphygmomanometers. The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated. The following parts are necessary for the NIBP measurement: • • • hemo module twin hose (adult or infant model) blood pressure cuffs (various sizes) 4.4.2 ECG Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential produced with electrodes placed on the surface of the body. ECG reflects: • • • • electrical activity of the heart normal/abnormal function of the heart effects of anesthesia on heart function effects of surgery on heart function See the “User's Guide” or the “User’s Reference Manual” for electrodes’ positions and other information. 4.4.3 Pulse oximetry A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near infrared (about 940 nm) and the other in the red region (about 660 nm) of the light spectrum. These wavelengths are emitted by LEDs in the SpO2 probe, the light is transmitted through peripheral tissue and is finally detected by a PIN-diode opposite the LEDs in the probe. The pulse oximeter derives the oxygen saturation (SpO2) using an empirically determined relationship between the relative absorption at the two wavelengths and the arterial oxygen saturation SaO2. In order to measure the arterial saturation accurately, pulse oximeters use the component of light absorption giving variations synchronous with heart beat as primary information on the arterial saturation. A general limitation of pulse oximetry is that due to the use of only two wavelengths, only two hemoglobin species can be discriminated by the measurement. The modern pulse oximeters are empirically calibrated either against fractional saturation SaO2frac; 1-38 Document no. 2081505-001 Hemo-dynamic module introduction HbO2 HbO2 + Hb + Dyshemoglo bin SaO2 frac = Formula 1 or against functional saturation SaO2func; SaO 2 func = HbO 2 HbO 2 + Hb Formula 2 Functional saturation is more insensitive to changes of carboxyhemoglobin and methemoglobin concentrations in blood. The oxygen saturation percentage SpO2 measured by the Datex-Ohmeda module is calibrated against functional saturation SaO2func. The advantage of this method is that the accuracy of SpO2 measurement relative to SaO2func can be maintained even at rather high concentrations of carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are not able to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or methemoglobin levels. Plethysmographic pulse wave The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at the measuring site. Thus the amplitude of the waveform represents the perfusion. Pulse rate The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals are filtered to reduce noise and checked to separate artifacts. Intensity of transmitted light Imax (DC-component) Imax Imin AC-component Variable absorption due to pulse added volume of arterial blood Transmitted light Arterial blood Tissue No pulsation Pulsatile blood Time absorption_of_light.vsd Venous blood Incident light Figure 17 Absorption of infrared light in the finger 1-39 Document no. 2081505-001 B40/B20 Patient Monitor SpO2 sensor connector RED Detector Figure 18 PSM_absorption_of_infrared.vsd IRED Emitter Pulse oximetry probe parts layout and schematic diagram The standard probe is a finger clamp probe which contains the light source LEDs in one half and the photodiode detector in the other half. Different kinds of probes are available from GE Healthcare. Summary of clinical studies used to establish the SpO2 accuracy claims NOTE: The pulse oximeter equipment measurements are statistically distributed, only about two-thirds of pulse oximeter equipment measurements can be expected to fall within ±Arms of the value measured by a CO-oximeter. The CO-oximeter should have a SaO2 accuracy of 1% (1 standard deviation) or better over the range of which the manufacturer makes SpO2 accuracy claims. A controlled desaturation study supporting a claimed range of SpO2 accuracy from 70 to 100% SaO2 can be supported with SaO2 data collected over the range of 73 to 97% SaO2. Nellcor Oximax Technology with Oximax sensors Measurement modules using Nellcor Oximax Technology with Oximax sensors have been validated for no motion accuracy in controlled hypoxia studies with healthy nonsmoking adult volunteers over the specified saturation SpO2 range(s). Pulse oximeter SpO2 readings were compared to SaO2 values of drawn blood samples measured by hemoximetry. Subjects used to validate SpO2 measurement accuracies were healthy and recruited from the local population. Subjects comprised both adult men and women and spanned a range of skin pigmentations. Because scatter and bias of pulse oximeter SpO2 and blood SaO2 comparisons commonly increase as the saturation decreases, and accuracy specifications are calculated from data spanning the stated range, different accuracy values may result when describing partially overlapping ranges. When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased by ± 1 digit, as compared to adult usage, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood. Masimo SET Technology with LNOP and LNCS sensors Measurement modules using Masimo SET Technology with LNOP and LNCS sensors have been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. Subjects comprised both adult men and women and spanned a range of skin pigmentations. One percent was added to accuracies for neonatal/infant sensors to account for accuracy variation due to properties of fetal hemoglobin. Refer to the directions for use of the Masimo SET sensors. 1-40 Document no. 2081505-001 Hemo-dynamic module introduction GE Trusignal Technology sensors Measurement modules using GE Trusignal Technology with OxyTip+/Trusignal sensors have been validated for no motion accuracy in controlled hypoxia studies with healthy nonsmoking adult volunteers over the specified saturation SpO2 range(s). Pulse oximeter SpO2 readings were compared to SaO2 values of drawn blood samples measured by hemoximetry. Subjects comprised both adult men and women and spanned a range of skin pigmentations. GE Oxy-AF and GE Oxy-SE sensors have been validated for neonatal accuracy. The subject demographics included 28 neonatals and 1 infant (15 females and 14 males). The subjects ranged in age from newborn to 37 days old. The weights ranged from 560 to 3060 g. The skin tones included in the study were light to dark. For neonatal study, the Arms of the collected convenience samples are 2.7, Oxy-AF sensor in the SaO2 range of 87-100% collected 52 data points, Oxy-SE sensor in the SaO2 range of 81-100% collected 53 data points. Test methods used to establish the accuracy claims during low perfusion Nellcor Oximax Technology with Oximax sensors Nellcor Oximax Technology with Oximax sensors have been validated for SpO2 low perfusion accuracy in bench top testing using Nellcor's PS II simulator with signal strength setting of 0.03% modulation and oxygen saturation levels of 70 to 100% at a pulse rate of 90 BPM. Nellcor Oximax Technology with Oximax sensors have been validated for low perfusion pulse rate accuracy in bench top testing using Nellcor's PS II simulator at a simulated low perfusion level of 0.10% modulation in the pulse rate range of 40 to 250 beats per minutes (BPM) at a SpO2 of 95%. Masimo SET Technology with LNOP and LNCS sensors Masimo SET Technology with LNOP and LNCS sensors have been validated for low perfusion accuracy in bench top testing against Biotek Index 2 Simulator and Masimo's simulator with signal strength setting of greater than 0.02% and a % transmission of greater than 5% for saturation ranging from 70%-100%. One percent was added to accuracies for neonatal/infant sensors to account for accuracy variation due to properties of fetal hemoglobin. Test methods used to establish the accuracy claims during motion Masimo SET Technology with LNOP and LNCS sensors Masimo SET Technology with LNOP and LNCS sensors have been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in the range of 70% to 100% SpO2 compared against a laboratory COoximeter and ECG monitor. The variation equals plus or minus one standard deviation, which encompasses 68% of the population. One percent was added to accuracies for neonatal/infant sensors to account for accuracy variation due to properties of fetal hemoglobin. NOTE: Accuracy during motion has not been specified for Masimo SET sensors LNOP TC-I, LNCS TC-I and LNCS TF-I. GE Trusignal Technology sensors Measurement modules using GE Trusignal Technology with TS-AF and TS-AP sensors have been validated for motion accuracy in controlled hypoxia studies with healthy non-smoking adult volunteers over the specified saturation SpO2 range(s). The following motion types were used: mechanically induced 3 Hz tapping motion at an amplitude of 1-2 cm, patient induced non-repetitive rubbing motion, and patient induced non-repetitive hand motion in supine 1-41 Document no. 2081505-001 B40/B20 Patient Monitor position. Pulse oximeter Sp02 readings were compared to SaO2 values of drawn blood samples measured by CO-oximetry. Subjects comprised both adult men and women and spanned a range of skin pigmentations. Test methods used to establish pulse rate accuracy Nellcor Oximax Technology with Oximax sensors Nellcor Oximax Technology with Oximax sensors have been validated for pulse rate accuracy in bench top testing using Nellcor's PS II simulator over the specified range at a SpO2 of 95%. Masimo SET Technology with LNOP and LNCS sensors Masimo SET Technology with LNOP and LNCS sensors have been validated for pulse rate accuracy over the specified range in bench top testing against a Biotek Index 2 simulator. GE Trusignal Technology sensors Measurement modules using GE Trusignal technology sensors have been validated for pulse rate accuracy over the specified range in bench top testing against a patient simulator. 4.4.4 Temperature The temperature is measured by a probe whose resistance varies when the temperature changes, called NTC (Negative Temperature Coefficient) resistor. The resistance can be measured by two complementary methods: • Applying a constant voltage across the resistor and measuring the current that flows through it. • Applying a constant current through the resistor and measuring the voltage that is generated across it. Hemo module uses the constant current method. The NTC-resistor is connected in series with a normal resistor and a constant current is applied through them. The temperature dependent voltage can be detected at the junction of the resistors, thus producing the temperature signal from the patient. The signal is amplified by analog amplifiers and further processed by digital electronics. 4.4.5 Invasive blood pressure To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of a connecting tubing, a pressure transducer, an intravenous bag of normal saline, all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed. The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed. 4.4.6 Respiration Impedance respiration is measured across the thorax between ECG electrodes. The respiration signal is made by supplying current between the electrodes and by measuring the differential current from the electrodes. The signal measured is the impedance change caused by breathing. The respiration rate is calculated from these impedance changes, and the respiration waveform is displayed on the screen. 1-42 Document no. 2081505-001 Hemo-dynamic module introduction Description of test method, data rate and method of ET-calculation The module uses concentration waveforms with data rate of 25Hz to calculate end-tidal (ET) gas readings. The module finds the time instant of the highest CO2 concentration in breath. Concentration at that instant is the ET CO2 reading. Because nitrous oxide and anesthetc agents are measured by the same sensor as CO2, the ET-readings of those gases are obtained directly at the time instant of ET CO2. For calculating ET-reading of O2 is then determined as O2-concentration at the time instant of ET CO2. If no breaths are detected for a given time (20s, for example), an apnea situation is triggered. During apnea, the ET values are updated every two seconds to the current concentration of each gas. The rated respiration rate range and the corresponding end-tidal gas reading accuracy were tested with reference gases of known concentrations. The test gases ware fed to the gas sampling system of the module through on electrically actuated valve with very low internal volume. Depending on its actuation status, the valve directed either room air or a test gas to the gas sampling line. The desired respiration rates were set by the electrical actuating times of the valve. The measurement accuracy of the end-tidal gas readings was tested using gas sampling lines of 3 meter length, connected to the gas sample port on the D-fend Pro water trap. The gas sampled to the sampling line was switched from room air to the test gases using on electrically actuated valve with low internal dead space to generate step changes in the gas concentrations. The electric actuating signal of the valve was generated using a highly accurate singal generator to accurately control the simulated respiration rate. The electronic sampling rate of the gas sensor signals is 25Hz, equaling a new data point on the gas waveform traces every 40 milliseconds. 1-43 Document no. 2081505-001 B40/B20 Patient Monitor 1-44 Document no. 2081505-001 2 Hardware installation Installation 1 Installation Safety precautions Warnings • The monitor or its components should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the monitor and its components should be observed to verify normal operation in the configuration in which it will be used. • After transferring or reinstalling the monitor, always check that it is properly connected and all parts are securely attached. Pay special attention to this in case of stacked mounting. • • • Use only manufacturer approved mounts. • EXPLOSION OR FIRE -Using non-recommended batteries could result in injury/burns to the patients or users. Only use batteries recommended or manufactured by GE. • Do not incinerate a battery or store at high temperatures. Serious injury or death could result. • Allow temperature of the monitor to stabilize to operating temperature range before connecting to power line. • NETWORK INTEGRITY — The monitor resides on the CARESCAPE, S/5 network. It is possible that inadvertent or malicious network activity could adversely affect patient monitoring. The integrity of the network is the responsibility of the hospital. • INTERFACING OTHER EQUIPMENT - Devices may only be interconnected with each other or to parts of the system when it has been determined by qualified biomedical personnel that there is no danger to the patient, the operator, or the environment as a result. In those instances where there is any element of doubt concerning the safety of the connected devices, the user must contact the manufacturers concerned (or other informed experts) for proper use. In all cases, safe and proper operation should be verified with the applicable manufacturer’s instructions for use, and system standards IEC 60601-1 must be complied with. • The medical electrical equipment or medical electrical system may be interfered with by other equipment, even if that other equipment complies with CISPR EMISSION requirements. Don’t install the monitor above the patient. Do not use the monitor in high electromagnetic fields (for example, during magnetic resonance imaging). Cautions • The monitor display is fragile. Ensure that it is not placed near a heat source or exposed to mechanical shocks, pressure, moisture or direct sunlight. 2-1 Document no. 2081505-001 B40/B20 Patient Monitor 1.1 Unpacking instructions 1. Confirm that the packing box is undamaged. If the box is damaged, contact the shipper. 2. Open the top of the box and carefully unpack all components. 3. Confirm that all components are undamaged. If any of the components is damaged, contact the shipper. 4. Confirm that all components are included. If any of the components is missing, contact your GE Healthcare distributor. 1.2 Choosing location Environmental requirements • Install the patient monitor to a location that meets the specified environmental requirements of operating temperature, humidity and atmospheric pressure. • Set up the device in a location which affords sufficient ventilation. The ventilation openings of the device must not be obstructed. EMI & RFI interference: • The patient monitor should be isolated from sources of strong electromagnetic and radio frequency interference. NOTE: Refer to "ElectroMagnetic Compatibility" for more information. 1.3 Mounting the monitor Mounting of monitor to the Wall Mount, Rollstand, Wall Mount with standard arm or Counter Top Mount is described in a separate instruction sheet delivered with each mount. NOTE: Refer to the “Supplier and accessories” to identify the compatible mounting hardware. 1.4 Connection to power Testing the battery charge Before installing a battery to the patient monitor, verify the battery’s state of charge. Press the green TEST button on the battery. The number of charge level indicator LEDs that illuminate indicates the approximate charge remaining in the battery. − Four LEDs illuminated: 75% – 100% of full-charge capacity. − Three LEDs illuminated: 50% – 74.9% of full-charge capacity. − Two LEDs illuminated: 25% – 49.9% of full-charge capacity. − One LED illuminated: 10% – 24.9% of full-charge capacity. − One LED flashing: < 10% of full-charge capacity remaining. 2-2 Document no. 2081505-001 Installation Installing the batteries 1. Open the lid of the battery compartment by the 2. Put in the new battery. Make sure that the side of the monitor. Move the latch up or down. charging indicator is facing to the back side of the monitor, then push the battery in all the way, move the latch and close the lid. Power on the monitor, check the monitor indicators. Connecting to mains Connect the power cord to the mains power inlet at the back of the monitor and to the wall socket. NOTE: Before taking the monitor into use for the first time, the batteries should be fully charged. Keep the monitor connected to the mains until the Battery charging symbol disappears (may take up to 5 hours if the batteries are fully discharged). Battery charging symbol WARNING The power cord may only be connected to a three-wire, grounded, hospital grade receptacle 1.5 Check configuration Check the monitor’s license and record in B. Installation and checkout form, B40/B20. The later set up and check out shall according to monitor’s configuration. 1. Turn on the monitor, enter to the service menu: Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) 2. Enter to license control menu: SW Management > License 3. Record the product license. 4. If the configuration is different with your order, contact GE Healthcare distributor. 2-3 Document no. 2081505-001 B40/B20 Patient Monitor 1.6 Connection to Network WARNING WARNING CAUTION Do not use with iCentral software V5.0.3 and earlier. Do not use Mobile Care Server software V5.2 and earlier. Install HL7 network interfaces as specified, and only by qualified personnel. NOTE: Not all the features mentioned following is available you monitor, according to your monitor’s configuration to set up. The monitor has been verified to be able to work in CARESCAPETM Network and S/5 network environments. Other network infrastructures are not supported. The monitor has EMR connectivity. There are three ways to acquire trended vital sign data from patient monitor: HL7 directly from monitor, HL7 from the CARESCAPE Gateway, or connecting to the serial port of each monitor. The monitor HL7 (Health Level Seven) message match with IHE PCD-01 OBR/OBX format. • In CARESCAPE network, the monitor can: − communicate with GE CARESCAPE Central Station V1 − communicate with GE CARESCAPE pro CIC version 4.0.8, 4.1.1 and 5.1.0 − compatible with Aware Gateway Server V1.6 − compatible with CARESCAPE Gateway server V1 − compatible with Mobile Care Server V6.0 − compatible with S/5 Collect V4.0 The monitor can talk to at most 4 CIC Central stations/CARESCAPE Central Station, 1 Aware Gateway Server or 1 CARESCAPE Gateway Server, 1 Mobile Care Server and 1000 other devices simultaneously in one CARESCAPE network. The B40 monitor can’t be set as the Time Master in CARESCAPE network. • In S/5 network, the monitor can: − communication with iCentral version 5.1.1 and 5.1.2. − compatible with S/5 Collect V4.0 − compatible with Mobile Care Server later than V5.2 The monitor does not support Patient Data Server; B40 realtime patient data can't be viewed on other monitors (e.g. Dash 3000/4000/5000, Solar 8000, B850, B650). 2-4 Document no. 2081505-001 Installation CARESCAPE Network CIS/HIS Systems CARESCAPE Administrative VLAN Network 1 Hospital Intranet 1 Internet CARESCAPE IX VLAN Network CARESCAPE? CIC Pro CARESCAPE Monitor B850 CARESCAPE Monitor B650 Mobile Care Server CARESCAPE MC VLAN Network Layer 3 Switch 1 RX Network 1 HL7 PCD-01 Outbound HL7 PCD-01 Outbound Dash® Patient Monitor 1. B40/B20 Patient Monitor B40/B20 Patient Monitor HL7 outbound from monitor though CNI V2 switch, need ‘Unity and HL7’ license Hospital Network CIS/HIS Systems 2 Hospital Intranet 2 Internet Hospital Switch HL7 PCD-01 Outbound B40/B20 Patient Monitor 2. HL7 outbound from monitor though hospital switch, need ‘HL7’ license Notes • • This switch should support intervlan routing function. • If hospital or the third party design networks by themselves, the hospital or the third party has full responsibility for the function and the risk of their network. The hospital or the third party can choose GE offered switch or other types. If you need to use GE offered switch, please contact to GE ND&I Team to setup CARESCAPE Network Infrastucture version 2. 2-5 Document no. 2081505-001 B40/B20 Patient Monitor 1.6.1 Pre-installation requirements Ensure that the applicable network infrastructure is in place prior to the installation of the patient monitor. Acquire the network configuration information from the hospital IT or the related project documentation and installation files. MC Network • The MC Network infrastructure shall be installed according to the "CARESCAPE Network Configuration Guide". • The installation site of the patient monitor shall have a wall jack and a network patch cable for the MC Network. S/5 Network • The S/5 Network shall be installed according to the "S/5 Network Installation Guide". Refer to the "iCentral and iCentral Client Service Manual" for iCentral installation instructions. • The installation site of the patient monitor shall have a wall jack and a network patch cable for the S/5 Network. HL7 network The HL7 Network shall be installed according to the "B40/B20 patient monitor HL7 reference manual". 1.6.2 To connect the network MC and S/5 network Use the CAT-5 network cable to connect the monitor to the network. 1. Make sure that the power is switched off. 2. Connect the one RJ-45 connector to network port at the back of the monitor. 3. Connect the other RJ-45 connector to the corresponding port on the wallbox. 4. Turn on the monitor and set up the network configuraiton (refer to 1.6.3. Network configuration). 5. Confirm that the network symbol and 'Network made' message are displayed in the upper part of the screen. 1.6.3 Network configuration How to set up MC network 1. Connect network cable to the ethernet port on the rear of the monitor. Connect the other end of the cable to the clinical network. 2. Enter the network service mode in the monitor: Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Frame > Network 3. Set up Unity configuration, enter to Network Config • 2-6 Document no. 2081505-001 Unit Name: It is used for setting the unit name in the monitor. The default unit name is “X”. NOTE: Unit Name here should be set to be same as CIC Unit Name. Installation 4. WARNING • Bed Name: It is used for setting the bed name in the monitor. The default bed name is the last five characters of the MAC address, excluding the colon delimiter characters. • MCS IP Address It is used for setting Mobile Care Server’s IP address which the monitor will talk to. It is gray when network license is not “Unity” or “Unity+HL7”. • Save Changes: It is gray if the changes have not been done to the menu information. NOTE: When save changes the monitor need restart. Set up the TCP/IP configuration: TCP/IP Config > TCP/IP Config • Configure the monitor’s IP address, subnet mask, default gateway. NOTE: If the IP address or both of unit name and bed name duplicate with other CARESCAPE network devices, the alarm will be given to the new added monitors. • Save the configuration after configuring the TCP/IP. NOTE: After saving changes the monitor need restart. If only B40 monitors and Dash monitors in the Unity network, do not set up B40 monitors with the highest IP address. For this may cause the Unity time sync function can’t work with Dash monitors. 5. If the monitor does not connect to network correctly, you may also need to set up the speed and duplex configuration: TCP/IP Config > TCP/IP Config > Speed and Duplex Choose the suitable settings according to your network environment. The default setting is AUTO. How to set up S/5 network 1. Connect network cable to the ethernet port on the rear of the monitor. Connect the other end of the cable to the clinical network. 2. Enter the network service mode in the monitor: Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Frame > Network 3. Set up S/5 network configuration, enter to Dri config Set up following information: • DRI level: It is used for setting the monitor’s network communication. should be selected to 2009 • Virtual Plug ID: Should enter 5 digit numbers, which is the same as in iCentral. NOTE: If you don’t have the virtual plug ID number in iCentral, please refer to the “iCentral User's Reference Manual” to set up in iCentral first. NOTE: After saving changes the monitor need restart. • Transfer Mode: It is used for choosing transfer mode: DRI/ETH or DRI/UDP Should be selected to DRI/ETH • Care Area: It is used for setting care area “OR” or “Non OR” for monitor. The monitor with different care area settings will behave differently in iCentral. NOTE: After saving changes the monitor need restart. 2-7 Document no. 2081505-001 B40/B20 Patient Monitor • Save Changes: Select “Save Changes“ to take in use changes made in this page. NOTE: In S/5 network protocol, no need to set up the TCP/IP settings. NOTE: For other the network menu’s instruction, please refer to 3.2. Network in “Service Menu” chapter for more details. How to set up the HL7 1. Set up Unit name and bed name first: Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Frame > Network > Network Config > Unit name and Bed Name. 2. Enter the HL7 service mode in the monitor: Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Frame > Network > HL7. 3. Set up HL7 configuration, enter to HL7 Config Set up following information: • HL7 Receiver IP Set up the HL7 receiver IP address • HL7 Receiver Port Set up the HL7 receiver port • HL7 Interval Set up the interval value • HL7 Patient Class Set up HL7 patient class E: Emergency, I: Inpatient, O: Outpatient, P: Preadmit, R: Reocurring patient, B: Obstetrics, U: Unknown • Save Changes It is gray if the changes have not been done to the menu information. 1.7 Inserting and removing the E modules To use the E module, your monitor need pre-configure the extension rack from manufacture. The E module options: airway gas module (E-miniC, E-sCO, E-sCAiO and N-CAiO), E-Entropy module. 2-8 Document no. 2081505-001 Installation To insert module: 1. Align the module with the insertion guides 2. Push the module into the monitor frame until it clicks and stops. 3. Pull the module outwards to insure the module is firmly seated. To remove module: WARNING 1.8 1. Pressing the release latch, on the bottom of the module. 2. Grasp the module firmly and pull out of the Frame. Make sure not to drop it when it comes out. When detaching modules, be careful not to drop them. Always support with one hand while pulling out with other. Monitor connections 2 1 11 10 9 5 4 8 7 Figure 3 3 6 B40 Front view (1) Transportation handle (2) Alarm light (3) The Trim Knob (4) Keyboard/membrane switch (5) Battery compartment (6) Guide rail for GCX mounting (7) Mains power and battery LEDs (8) On/Off key (9) Hemo connectors B20 Front view (10) E modules (11) Recorder module 2-9 Document no. 2081505-001 B40/B20 Patient Monitor Figure 4 B40 back B20 back 1 4 6 5 7 2 3 Figure 5 I/O and connectors (1) Receptacle for power cord (2) Serial port (3) Defibrillator connector (4) Nurse call connector (5) Network connector (6) Equipotential connector (7) Multi I/O connector NOTE: The Multi I/O with ports 2,3,4 are optional parts for customer. 2-10 Document no. 2081505-001 Installation 1.9 Visual indicators Function Specification Explanation External power supply Green LED Indicates when monitor is powered from mains Battery operation Green LED Indicates when monitor is powered from internal batteries Battery condition Orange LED Indicates when monitor is charging batteries (solid) or battery failure (flashing). Alarm Light Highly visible Red/Yellow/Cyan light Ease alarm detection from distance. 1.10 Installation checkout Refer to the 3. Installation checkout in Chapter 3 for procedure. 2-11 Document no. 2081505-001 B40/B20 Patient Monitor 2-12 Document no. 2081505-001 3 Maintenance Instructions 1 Instructions 1.1 Introduction These instructions include procedures of system maintenance for the B40. It’s include four sections: • • Electrical safety tests. • • Maintenance and checkout, which should be performed every 24 months. Installation checkout, which should be performed after installation and service configuration. Adjustments and calibrations NOTE: Please complete the check form when performing the corresponding procedures. NOTE: For the Gas modules maintenance and calibration, please refer to “7. E-miniC Module” and “8. E-sCO, E-sCAiO and N-CAiO Modules“. NOTE: For the Entropy module maintenance and calibration, please refer to “9. E-Entropy Module“. The symbol in the instructions means that the procedure performed should be signed in the check form. The procedures should be performed in ascending order, bypassing those that are not applicable for a particular monitor. To enter the service menus, you need following passwords: Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) In case you evaluate the measurement accuracy with a patient simulator, add the simulator’s accuracy specification to the one for the monitor. CAUTION CAUTION CAUTION Failure on the part of all responsible individuals, hospitals or institutions, employing the use of this device, to implement the recommended maintenance schedule may cause equipment failure. The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule, unless an equipment maintenance agreement exists. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device. Only trained personnel with appropriate equipment should perform the tests and repairs outlined in this section. Unauthorized service may void warranty of the unit. Wear a static control wrist strap when handling PCB boards. Electrostatic discharge may damage components on the board. 3-1 Document no. 2081505-001 B40/B20 Patient Monitor 1.2 Recommended tools NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy. NOTE: A functional tester cannot be used to assess the accuracy of pulse oximeter for monitor. Table 1 Recommended accessories and tools Accessories A rigid cylinder or pipe NIBP cuff Adult NIBP cuff hose with cuff ID Infant NIBP cuff hose with cuff ID Tubing parts to connect a manometer and a pump to the NIPB cuff and hose. Dual invasive pressure adapter cable ECG accessories, IEC or AHA - Multi-link 3-lead integrated cable and leadwire - Multi-link 5-leadwire set - Multi-link 3/5-lead ECG trunk cable SpO2 finger probe SpO2 Interconnect Cable Temperature dual cable CO2 Sampling line 3m/10 ft Tool A multiparameter patient simulator with IBP, Temp adpter cables Screwdrivers PH1, PH2 NOTE: For details on recommended accessories see “Supplies and Accessories“ catalog. 3-2 Document no. 2081505-001 Electrical Safety Tests 2 Electrical Safety Tests Electrical safety tests provide a method of determining if potential electrical health hazards to the patient or operator of the device exist. GE recommends that the qualified personnel performing the tests. 2.1 Test setup Test conditions Perform electrical safety tests under normal ambient conditions of temperature, humidity and pressure. Test equipment The recommended test equipment required to perform electrical safety tests is listed below. Tool Specification Satety Analyzer/Leakage Current Tester Equivalent to the circuits shown Ground Bond Tester 0 – 1 ohm Safety Test Body Kita P/N M1155870 or equivalent a Instead of the test bodies included in the safety test body kit, other applicable test bodies with all pins connected together may be used. Perform electrical safety tests using an electrical safety analyzer per IEC 60601-1, UL 60601-1, EN 60601-1 or CSA C22.2 No. 601.1. The schematics in the section provide a general understanding of the test equipment. Actual configuration of test equipment may vary. The patient monitor being tested should be placed on an insulating surface. NOTE: Before proceeding, make sure that all test equipment is properly calibrated, maintained and functioning. NOTE: Refer to the instructions delivered with the safety analyzer to perform each test. NOTE: GE recommends that the qualified personnel performing the tests should record the test results of each electrical safety test, for example by using the installation / maintenance check forms included in this manual. System setup These instructions are intended for every component in the system. Ensure that all system components are properly connected to the patient monitor as described in chapter “2. Hardware installation”. 2.2 Power Outlet Test Verify that the power outlet is wired correctly per the country’s electrical code standard before starting the following electrical safety tests. The results of the following tests will be inaccurate unless a properly wired power outlet is used. Use only non-isolated power outlets when performing safety tests. 3-3 Document no. 2081505-001 B40/B20 Patient Monitor 2.3 Power cord and plug Verify the power cord being used with the patient monitor is good. The following are a couple of things to check for in this regard: • Inspect the power cord for wear or damage regularly. If damage is suspected, test for continuity through each conductor of the power cord connector. • Verify line, neutral, and earth conductors are properly connected to the power cord plug and are not short-circuited. Replace the power cord, as necessary with a regulatory-approved cord for the country of use. WARNING Use only AC power cords recommended or manufactured by GE. 2.4 Ground (earth) integrity Listed below are two methods for checking the ground (earth) integrity, “Ground Continuity Test” and “Impedance of Protective Earth Connection.” These tests determine whether the device's exposed metal and power inlet's earth (ground) connection has a power ground fault condition. Perform the in accordance with your local regulations. Refer to the instructions contained with the safety analyzer to perform each test. 2.4.1 Ground Continuity Test The measuring device (MD) in the diagram below may be a DMM or part of a safety analyzer. NOTE: The measuring device (MD) represents the network and voltage measuring instrument and its frequency characteristics per IEC 60601-1. 3-4 Document no. 2081505-001 Electrical Safety Tests Acceptance criteria: • For equipment without a power supply cord, the impedance between the protective earth terminal and any accessible metal part which is protectively earthed shall not exceed 0.1 ohms. • For equipment with a power supply cord, the impedance between the protective earth pin in the mains plug and any accessible metal part which is protectively earthed shall not exceed 0.2 ohms. 2.4.2 Impedance of Protective Earth Connection This test, unlike a ground continuity test, will also stress the ground system by using special ground bond testers. This test normally is only required as a manufacturing production test to receive safety agency compliance. Some country agencies do require this test after field equipment repairs (i.e., Germany’s DIN VDE 0751 standards). Consult your country/local safety agency if in question. Check compliance as follows: (1) A current of 25A from a current source with a frequency of 50 or 60 Hz with a no-load voltage not exceeding 6 V is passed for at least 5 seconds, but no more than 10 seconds, through the protective earth terminal or the protective earth pin in the mains plug and each accessible metal part which could become live in case of failure in basic insulation. (2) The voltage drop between the parts described is measured and the impedance determined from the current and voltage drop. It shall not exceed the values indicated. When taking this measurement, move the unit’s power cord around. There should be no fluctuations in resistance. Acceptance criteria: • For equipment without a power supply cord, the impedance between the protective earth terminal any accessible metal part which is protectively earthed shall not exceed 0.1 Ohms. • For equipment with a power supply cord, the impedance between the protective earth pin in the mains plug and any accessible metal part which is protectively earthed shall not exceed 0.2 ohms. 2.5 Earth leakage current test This test measures the current leakage flowing from the mains part through or across the insulation into the protective earth conductor of the device under test. Perform this test both in Normal Condition (NC) and in a Single Fault Condition (SFC), where one of the supply conductors is open at a time. Perform the test with normal and reverse polarity. NOTE: Refer to the instructions delivered with the safety analyzer to perform this test. 3-5 Document no. 2081505-001 B40/B20 Patient Monitor Power Cord Leakage Tester HIGH NORM Power Cord LOW GND GND RVS Device Under Test 0.15µF DMM 1K 10 DMM set to measure AC voltage NOTE: The measuring device (MD) represents the network and voltage measuring instrument and its frequency characteristics per IEC 60601-1. (1) Configure the safety analyzer as follows (NC): • • Polarity - NORMAL Neutral - CLOSED (2) Power on the device under test. (3) Read and record the current leakage indicated on the tester. (4) Configure the safety analyzer as follows (SFC): • • Polarity - NORMAL Neutral - OPEN (5) Read and record the current leakage indicated on the tester. (6) Configure the safety analyzer as follows (SFC): • • Polarity - REVERSED Neutral - OPEN (7) Read and record the current leakage indicated on the tester. (8) Configure the safety analyzer as follows (NC): • • (9) Polarity - REVERSED Neutral - CLOSED Read and record the current leakage indicated on the tester. (10) Power off the device under test. Acceptance criteria in NC (Normal condition): • All readings shall be less than or equal to 300 μA for installations that require compliance to UL 60601-1 requirements. • All readings shall be less than or equal to 500 μA for installations that require compliance to EN 60601-1 / IEC 60601-1 requirements. 3-6 Document no. 2081505-001 Electrical Safety Tests Acceptance criteria in SFC (Single fault condition) - one of the supply conductors open at a time: • All readings shall be less than or equal to 1 mA. 2.6 Enclosure (Touch) leakage current test This test measures current leakage through the exposed conductive parts on the device under test. Perform the test in Normal Condition (NC) and in two different Single Fault Conditions (SFC): 1) earth open and 2) one of the supply conductors open at a time. Perform the test with normal and reverse polarity. NOTE: Refer to the instructions delivered with the safety analyzer to perform this test. Power Cord Leakage Tester HIGH NORM Power Cord LOW GND RVS Open Closed GND Device Under Test 0.15µF DMM 1K 10 Probe to exposed conductive chassis DMM set to measure AC voltage NOTE: The MD represents the network and voltage measuring instrument and its frequency characteristics per IEC 60601-1. (1) Configure the safety analyzer as follows (NC): • • • Polarity - NORMAL Neutral - CLOSED GND (Earth) - CLOSED (2) Power on device under test. (3) Read and record the current leakage indicated on tester. (4) Configure the safety analyzer as follows (SFC): • • • Polarity - NORMAL Neutral - OPEN GND (Earth) - CLOSED (5) Read and record the current leakage indicated on the tester. (6) Configure the safety analyzer as follows (SFC): • • • Polarity - NORMAL Neutral - CLOSED GND (Earth) - OPEN 3-7 Document no. 2081505-001 B40/B20 Patient Monitor (7) Read and record the current leakage indicated on the tester. (8) Configure the safety analyzer as follows (SFC): • • • (9) Polarity - REVERSED Neutral - CLOSED GND (Earth) - OPEN Read and record the current leakage indicated on the tester. (10) Configure the safety analyzer as follows (SFC): • • • Polarity - REVERSED Neutral - OPEN GND (Earth) - CLOSED (11) Read and record the current leakage indicated on the tester. (12) Configure the safety analyzer as follows (NC): • • • Polarity - REVERSED Neutral - CLOSED GND (Earth) - CLOSED (13) Read and record the current leakage indicated on the tester. (14) Power off the device under test. Acceptance criteria in NC: • All readings shall be less than or equal to 100 µA Acceptance criteria in SFC - earth open or one of the supply conductors open at a time: • All readings shall be less than or equal to 300 μA for installations that require compliance to UL 60601-1 requirements. • All readings shall be less than or equal to 500 μA for installations that require compliance to EN 60601-1 / IEC 60601-1 requirements. 2.7 Patient leakage current tests - overview The following table specifies the parameter modules and the related patient connectors to be tested in the “2.7.1. Patient (source) leakage current test” and in the “2.7.2. Patient (sink) leakage current tests”. Use the safety test body kit, P/N M1155870 (or equivalent), to perform patient leakage current tests. This safety test body kit contains various patient connectors where all pins are shorted out together. For information on which test body to use for each patient connector, refer to the service instructions included in the safety test body kit. NOTE: If not otherwise stated in the table below, each test body is connected directly to the specified connector in the patient module. Table 2 3-8 Document no. 2081505-001 Patient connectors to be tested with each module Electrical Safety Tests Module Patient connector Hemo ECG and SpO2 E-Entropy 1. Connect an Entropy sensor cable to the module. 2. Connect the specified test body to the Entropy sensor and cable. 2.7.1 Patient (source) leakage current test This procedure measures the leakage current from an applied part connector of the device to ground. Perform the test in Normal Condition (NC) and in two different Single Fault Conditions (SFC): 1) earth open and 2) one of the supply conductors open at a time. Perform test with normal and reverse polarity. NOTE: Refer to the instructions delivered with the safety analyzer to perform this test. AC Mains Leakage Tester /,1( NORM Power Cord 1(875$/ ($57+ RVS Device Under Test 120K Closed GND 0.15μF DMM 1K Patient connector Test Body 10 NOTE: The MD represents the network and voltage measuring instrument and its frequency characteristics per IEC 60601-1. NOTE: Perform this test for all the connected parameter modules and patient connectors specified in Table 2. (1) Configure the safety analyzer as follows (NC): • • • Polarity - NORMAL Neutral - CLOSED GND (Earth) - CLOSED (2) Power on the device under test. (3) Read and record the current leakage indicated on the tester. (4) Configure the safety analyzer as follows (SFC): • • • Polarity - NORMAL Neutral - OPEN GND (Earth) - CLOSED 3-9 Document no. 2081505-001 B40/B20 Patient Monitor (5) Read and record the current leakage indicated on the tester. (6) Configure the safety analyzer as follows (SFC): • • • Polarity - NORMAL Neutral - CLOSED GND (Earth) - OPEN (7) Read and record the current leakage indicated on the tester. (8) Configure the safety analyzer as follows (SFC): • • • (9) Polarity - REVERSED Neutral - CLOSED GND (Earth) - OPEN Read and record the current leakage indicated on the tester. (10) Configure the safety analyzer as follows (SFC): • • • Polarity - REVERSED Neutral - OPEN GND (Earth) - CLOSED (11) Read and record the current leakage indicated on the tester. (12) Configure the safety analyzer as follows (NC): • • • Polarity - REVERSED Neutral - CLOSED GND (Earth) - CLOSED (13) Read and record the current leakage indicated on the tester. (14) Power off the device under test. (15) Repeat this test for all the connected parameter modules and patient connectors specified in Table 2. Acceptance criteria in NC: • All readings shall be less than or equal to 10 µA Acceptance criteria in SFC - earth open or one of the supply conductors open at a time • All readings shall be less than or equal to 50 µA 2.7.2 Patient (sink) leakage current tests This procedure measures the leakage current from an applied part connector of the device to ground. Perform the test in Normal Condition (NC) with normal and reverse polarity. NOTE: Refer to the instructions delivered with the safety analyzer to perform this test. 3-10 Document no. 2081505-001 Electrical Safety Tests Power Cord Leakage Tester LINE NORM Power Cord NEUTRAL EARTH RVS Device Under Test 120K Closed GND Patient connector Test Body 0.15μF DMM 1K 10 (Keep cable length as short as possible.) NOTE: The MD represents the network and voltage measuring instrument and its frequency characteristics per IEC 60601-1. NOTE: Per IEC 60601-1, the impedance to protect the circuitry and the person performing the test, but low enough to accept currents higher than the allowable values of the LEAKAGE CURRENT to be measured. WARNING Shock hazard. The following step causes high voltage at the test body. Do not touch the test body. NOTE: Perform this test for all the connected parameter modules and patient connectors specified in Table 2. (1) Configure the safety analyzer as follows: • • • Polarity - NORMAL Neutral - CLOSED GND (Earth) - CLOSED (2) Power on the device under test. (3) Read and record the current leakage indicated on the tester. (4) Configure the safety analyzer as follows: • • • Polarity - REVERSED Neutral - CLOSED GND (Earth) - CLOSED (5) Read and record the current leakage indicated on the tester. (6) Power off the device under test. (7) Repeat this test for all the connected parameter modules and patient connectors specified in Table 2. Acceptance criteria: • All readings shall be less than or equal to 50 µA 3-11 Document no. 2081505-001 B40/B20 Patient Monitor 2.8 Test completion (1) Disconnect the safety analyzer from the power outlet. (2) Disconnect the test equipment from the patient monitor. (3) Disconnect the patient monitor’s power cord from the leakage tester. (4) Fill in all necessary documents. 3-12 Document no. 2081505-001 Installation checkout 3 Installation checkout The purpose of the installation checkout procedure is to ensure that the system is properly installed and configured for use. Service personnel shall perform the following checkout procedure for the monitoring system after the hardware installation and service configuration is completed: 1. 3.1. Visual inspection 2. 3.2. Functional inspection Skip the tests that are not applicable for the installed monitor. These instructions include a "Installation and checkout form, B40/B20" on page B-1 to be filled in when performing the procedures. 3.1 Visual inspection Perform the following visual inspection to the installed monitoring system: • • • Carefully inspect the patient monitor for any damage. • • Verify that the modules are properly connected and locked in place. Verify that the patient monitor is properly mounted with specified mounting solutions. Verify that the cables between the patient monitor and the connected peripheral devices are intact and properly connected to the right connectors. Verify that the battery door is properly locked. The cleaning precautions, cleaning requirements, cleaning procedures, and recommended cleaning solutions for the monitor are described in the "User’s Guide". For details about cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the accessory package. 3.2 Functional inspection 3.2.1 Start-up 1. Turn on the patient monitor. Verify that the monitor starts up normally: • • • The red, yellow and cyan alarm lights are lit in sequence. • Check and there are no error messages on the screen. The speaker gives an audible beep. Check that the GE logo screen is displayed, followed by the notes screen and the normal monitoring screen appears. NOTE: Refer to section "Condition the battery" to see the procedure for battery conditioning if you receive a a Condition Battery X message. NOTE: Before taking the patient monitor into use for the first time, the battery should be fully charged. Keep the monitor connected to the mains until the battery is fully charged. 3-13 Document no. 2081505-001 B40/B20 Patient Monitor 3.2.2 Display 1. Verify that all text is readable and all images are clear. 2. Verify that the brightness is good. Adjust if necessary. 3.2.3 Time and date 1. Check that the clock on the screen shows correct time. Adjust the time and date, if necessary. Monitor Setup < Time and Date NOTE: The monitor can’t be set as the TIME MASTER in network. You should adjust the time and date from the central station. 3.2.4 Parameters measurements Connect the accessories (no need to connect simulator/patient), check the monitor displays the following messages or activities. • • ECG: After connecting ECG cable, ‘leads off’ will display in the Waveform Field • • SpO2: After connecting SpO2 cable and sensor, SpO2 sensor will be lit. NIBP: After connecting NIBP hose to module, ‘Adult/Pediatric’ or ‘Neonatal’ will display in NIBP Digital Field for several seconds Temperature: After connecting Temp cable and sensor, ‘Performing temp test:’ will display in Temp Digital Field for several seconds. • IBP: After connecting IBP cable and transducer, ‘InvBP’s not Zeroed’ will display in Message Field. • Gas: After installing the gas module, ‘Calibrating gas sensor’ will display in CO2 waveform field for about 1 minutes. • Entropy: After installing the E-Entropy module and cable, 'No sensor' message will display in Entropy digital field. 3.2.5 Recorder 1. Press the Recorder Start/Stop key and check that the module starts recording the selected waveforms. Press the Recorder Start/Stop key again to stop recording. 2. Check that the quality of the recordings is acceptable. 3.2.6 MC or S/5 Network connection NOTE: Pre-configure the network when install the monitor. 3-14 Document no. 2081505-001 Installation checkout 1. Check that the CAT-5 cable connector is clean and intact, then connect it to the Network connector on the backside of the monitor. Check that the monitor connects to the network, i.e. the network symbol appears on the upper right-hand corner of the screen. 3.2.7 Conclusion • • • Power off the monitor Perform final cleaning Fill in all necessary documents, refer to "Installation and checkout form, B40/B20" on page B-1 3-15 Document no. 2081505-001 B40/B20 Patient Monitor 4 Maintenance and checkout These instructions include procedures for maintenance and checkout for the B40. Service personnel shall perform the following checkout procedure every 24 months after installation: 1. 4.1. Visual inspection 2. 2. Electrical Safety Tests 3. 4.2. Functional inspection 4. 4.3. Monitor battery maintenance Skip the tests that are not applicable for the installed monitor. These instructions include "Maintenance and checkout form, B40/B20" on page C-1, which to be filled in when performing the corresponding procedures. Before starting • Save the patient data and monitor settings if necessary. • Make sure that the monitor is turned off. • Disconnect the mains power cord. • If the monitor is connected to the network, disconnect the CAT-5 cable from the monitor. 4.1 Visual inspection Perform the following visual inspection to the installed monitoring system: • • • Carefully inspect the patient monitor for any damage. • • Verify that the modules are properly connected and locked in place. Verify that the patient monitor is properly mounted with specified mounting solutions. Verify that the cables between the patient monitor and the connected peripheral devices are intact and properly connected to the right connectors. Verify that the battery door is properly locked. The cleaning precautions, cleaning requirements, cleaning procedures, and recommended cleaning solutions for the monitor are described in the "User’s Guide". For details about cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the accessory package. 4.2 Functional inspection 4.2.1 Start-up 1. Turn on the patient monitor. Verify that the monitor starts up normally: • • 3-16 Document no. 2081505-001 The red, yellow and cyan alarm lights are lit in sequence. The speaker gives an audible beep. Maintenance and checkout • Check that the GE logo screen is displayed, followed by the notes screen and the normal monitoring screen appears. • Check and there are no error messages on the screen. NOTE: Refer to section "Condition the battery" on page 3-23 to see the procedure for battery conditioning if you receive a a Condition Battery X message. NOTE: Batteries are recommended to be conditioned every six months. 4.2.2 Display 1. Verify that all text is readable and all images are clear. 2. Verify that the brightness is good. Adjust if necessary. 4.2.3 Keyboard(s) 1. Tests with the keyboard/membrane switch: − Enter the service menu: Monitor Setup < Install/Service (16-4-34) < Service (password 26-23-8) Keyboard − Check the keys one by one, turn trim knob one full turn clockwise and one full turn counter clockwise. 4.2.4 Time and date 1. Check that the clock on the screen shows correct time. Adjust the time and date, if necessary. Monitor Setup < Time and Date NOTE: The monitor can’t be set as the TIME MASTER in network. You should adjust the time and date from the central station. 4.2.5 Hemo Module ECG measurement • Configure ECG settings in monitor: Connect 5 lead ECG cable to monitor and simulator. Monitor Setup < Screen Setup < Waveform Fields < Field 1: ECG1; Field 2: ECG2; Field 3: ECG3 ECG < ECG1 Lead: II; ECG2 Lead: V1; ECG3 Lead: aVL; ECG Size: 1.0 < ECG Setup < Beat Sound Volume: 1 or greater; Pacemaker: Show; HR Source: AUTO • Configure ECG settings in simulator: ECG rhythm: a normal sinus rhythm 3-17 Document no. 2081505-001 B40/B20 Patient Monitor heart rate: 80 bpm Amplitude: 1 mV 1. Normal Sinus Rhythm − Check that the monitor displays the ECG leads II, V1 & aVL and the waveforms are noise-free. The monitor shall display a 80 ± 5 bpm heart rate and an audible QRS tone sounds with each QRS complex. 2. Pacemaker Detection − Configure the simulator to output "Asynchronous Pacemaker Pulse" − Check that pacemaker spikes are shown on the ECG waveform. 3. Asystole Detection − Configure the simulator to output "Asystole". − Check that the 'Asystole' alarm appears to the monitor screen. − Configure the simulator to show "80 beats per minute, Normal Sinus Rhythm". 4. Leads Off Detection − Detach the RA/R leadwire from the simulator. − Check that the Lead II waveform disappears from the ECG1 waveform field and a message 'RA/R lead off'' is shown momentarily. − Check that Lead II is replaced by Lead III in the ECG1 waveform field after a while and a message 'Lead changed' is followed by a message 'Learning'. − Reconnect the RA/R leadwire to the simulator. − Check that Lead III is replaced back to Lead II in the ECG waveform field. Respiration measurement • Configure RESP settings in monitor: Set up the Resp waveform field to the monitor screen: Monitor Setup < Screen Setup < Waveform Fields < Field 4: Resp And Others < RespSetup < Resp Rate Source: Imped. < Measurement: ON • Configure RESP settings in simulator: Baseline impedance: 1000 Amplitude: 1 Respiration rate: 20 breaths per minute Lead selection: II (or LL) 5. Respiration Rate − Check that the RESP waveform is shown and the RR value is 20 (±5). − Configure the simulator's Apnea Simulation to "32 sec". 3-18 Document no. 2081505-001 Maintenance and checkout 6. Apnea Detection − Check that the monitor activates the Apnea alarm. − Configure the simulator's Apnea Simulation to "OFF" Temperature measurement • Configure the “T1+T2” digit field to the monitor screen. • Configure the simulator’s temperature channels as follows: Temperature : 37 °C/98.6 °F 7. Temperature detection − Check the corresponding temperature value appears and that no error messages are shown on the monitor screen. NOTE: If the deviation on a temperature reading on the screen is more than 0.1°C, calibrate the temperature channels according to the instructions in chapter "Temperature calibration" on page 3-25. Invasive blood pressure measurement • Configure the simulator’s IBP channels as follows: Sensitivity: 5 µV/V/mmHg InvBP outputs: "0 mmHg static pressure" or "atmosphere" 8. Zeroing − Press IBP Zero All key. − Check that a message "Zeroing" followed by a message "Zeroed" is shown in the IBP parameter window. 9. Static Pressure − Configure the simulator's InvBP output to "200 mmHg static pressure". − Check that the flat pressure line appears on the related waveform field. The reading in the parameter window shall be 200 ±10 mmHg. NOTE: Recalibration is required, if the measured value is not within the specification. Calibrate the invasive pressure channels according to the instructions in "Invasive pressure calibration" on page 3-26. 10. Pressure Waveforms − Configure the simulator's InvBP output to "Arterial 120/80". − Check that the pressure waveform for tested invasive pressure channel appears in the IBP waveform field and the Sys/Dia (Mean) pressure values are shown in the related parameter window. 3-19 Document no. 2081505-001 B40/B20 Patient Monitor SpO2 measurement 11. Test measurement Connect the SpO2 probe onto your finger. Check that the reading of 90-100 and SpO2 waveform appears. Check that the HR value is calculated from SpO2 when ECG and IBP (ABP or Art) are not measured. NOTE: a functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a pulse oximeter monitor Non Invasive Blood Pressure measurement • Connect and set up the NIBP integrated tool as following picture, ensure all the connections made are leak-proof. 12. NIBP Leak Test − Enter the NIBP Calibration in service menu: Parameters < NIBP < Calibrations − Select Active Leak Test: ON − Wait for 15 seconds for the pressure to stabilize then check that the pressure does not drop more than 5 mmHg per one minute. 13. NIBP calibration − Calibrate the Non-invasive blood pressure (NIBP) channel according to the instructions in "NIBP calibrations" on page 3-24. 14. NIBP hose detection − 3-20 Document no. 2081505-001 Disconnect the calibration test equipment. Maintenance and checkout − Make sure NIBP Setup - Inflation Limits is set to Auto. − Attach an A/P NIBP cuff hose without cuff identification. − Press the NIBP Start/Cancel key. After a few moments the monitor automatically opens the selections NIBP Setup - Inflation Limits for user to manually select the inflation limits. 4.2.6 Loudspeaker Check the loudspeaker by setting the alarm volume: Alarms Setup < Alarm Volume Test the whole volume scale from 1 to 10 by turning the Trim Knob and check that the alarm volume changes correspondingly. The alarm sound should be clear and audible with all the settings. 4.2.7 Monitor software Enter the service menu: Monitor Setup < Install Service (16-4-34) < Service (26-23-8) Take down the information regarding Monitor software. 4.2.8 Watchdog circuitry Select: Monitor Setup < Install Service (16-4-34) < Service (26-23-8) < Set/Test Perform the tests Watchdog and WD by Overload. Check that the monitor restarts in each case. NOTE: When selecting Watchdog and WD by Overload, auto restarting should take place approximately after 10 seconds. 4.2.9 Network NOTE: Pre-configure the network when install the monitor. 1. Check that the CAT-5 cable connector is clean and intact, then connect it to the Network connector on the backside of the monitor. Check that the monitor connects to the network, i.e. the network symbol appears on the upper right-hand corner of the screen. 3-21 Document no. 2081505-001 B40/B20 Patient Monitor 4.2.10 Final cleaning Switch off the monitor and perform final cleaning. Fill in all necessary documents, refer to "Maintenance and checkout form, B40/B20" on page C-1. 4.3 Monitor battery maintenance The lithium-ion (Li-Ion) battery is a rechargeable battery containing lithium-ion cells. Each battery contains an integrated electronic fuel gauge and a safety protection circuit. The following are facts about lithium-ion battery technology: • The battery discharges on its own, even when it is not installed in the equipment. This discharge is the result of the lithium-ion cells and the bias current required for the integrated electronics. • • The capacity loss of the battery degrades significantly at higher temperatures. As the battery ages, the full-charge capacity of the battery degrades and is permanently lost. As a result, the amount of charge that is stored and available for use is reduced. 4.3.1 Use recommendations GE recommends the following methods to improve battery performance: − Location — Position the equipment in a location that does not artificially increase the operating temperature of the batteries. − Conditioning guideline — Condition the battery when a ‘Condition Battery X’ message is shown on the monitor screen. The condition cycle recalibrates the electronic fuel gauge. 4.3.2 Storage recommendations GE recommends storing the battery outside of the device at a temperature between 20°C to 25°C (68°F to 77°F). 4.3.3 Test the battery charge Before installing a battery, verify the battery’s state of charge. Press the green TEST button on the battery. The number of charge level indicator LEDs that illuminate indicates the approximate charge remaining in the battery. − Four LEDs illuminated: 75% – 100% of full-charge capacity. − Three LEDs illuminated: 50% – 74.9% of full-charge capacity. − Two LEDs illuminated: 25% – 49.9% of full-charge capacity. − One LED illuminated: 10% – 24.9% of full-charge capacity. − One LED flashing: < 10% of full-charge capacity remaining. 4.3.4 Charge the battery The battery charges whenever it is installed into the patient monitor and the patient monitor is connected to an AC power source. The battery is charging both when the patient monitor is turned on and turn off. Battery is charging as long as the orange battery charging indicator LED is lit. 3-22 Document no. 2081505-001 Maintenance and checkout 4.3.5 Condition the battery Battery conditioning is needed if the ‘Condition Battery X’ message is displayed. Condition the battery by fully discharging and recharging the battery twice according to the following procedure: NOTE: The patient monitor must be in a discharged state during battery conditioning. Disconnect any acquisition modules from the patient monitor if connected. 1. Turn on the patient monitor. Disconnect the power cord from the wall outlet. Leave the patient monitor on until the battery is fully discharged and the patient monitor turns off automatically. NOTE: Ignore the ‘Battery Low’ and ‘Battery Empty’ messages when discharging the battery. 2. Reconnect the power cord to the wall outlet and turn on the patient monitor. Leave the patient monitor on until the battery is fully recharged and the orange battery charging indicator LED turns off. 3. Repeat the steps 1 and 2 once. The battery is now conditioned and ready for use. However if the ‘Condition Battery X’ message is still shown on the screen, replace the battery. 4.3.6 Replace the battery Replace the battery in the following situations: − If the ‘Replace Battery X’ message is displayed. This message indicates that the full-charge capacity of the used battery has considerably degraded compared to the design capacity of a new battery. Remove the battery from the monitor and install a new battery according to the procedure in section "Installing the batteries" on page 2-3. NOTE: Dispose of the battery according to local, state or country laws. WARNING Do not incinerate the battery or store at high temperatures. NOTE: Refer to section "Battery indicators" on page 4-6 for more detailed information about the battery status. 3-23 Document no. 2081505-001 B40/B20 Patient Monitor 5 Adjustments and calibrations NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy. NOTE: Discharge the patient before do the following procedures. 5.1 NIBP calibrations NIBP calibration should be performed: • • • Each time planned maintenance is performed. Each time corrective maintenance is performed. If the measured value is not within the specification. Calibration check • Refer to the "Non Invasive Blood Pressure measurement" on page 3-20 in maintenance checkout section for NIBP integrated tool set up. 1. Enter Calibration menu: Monitor Setup - Install/Service (password 16-4-34) -Service (26-23-8) - Parameters NIBP - Calibrations Calibration Active Leak Test OFF Calibration Check OFF Protection OFF Calibrate Previous Menu 2. Select Calibration Check and push the Trim Knob. 3. Connect an external precision manometer to the monitor, the same as "Non Invasive Blood Pressure measurement" on page 3-20. 4. Pump the following pressures to manometer and check the difference between the manometer and monitor pressure display (The zeroing offset is automatically subtracted from the pressure readings). Table 3 NIBP calibration check pressures Pressure Max. error Example 0 mmHg ±5 mmHg (=zero offset) -1 100 mmHg 100 ±2 mmHg 100 ±2 200 mmHg 200 ±3 mmHg 200 ±3 3-24 Document no. 2081505-001 Adjustments and calibrations There is a reading of B1 and B2 in help filed under NIBP calibration menu. If the error of pressure channel B1 and B2 is larger than specified above, the module should be recalibrated. The error of B2 has no effect on blood pressure measurement accuracy. NOTE: If the monitor pass the calibration check, no need to do NIBP calibration. Calibration 1. Enter Calibration menu. 2. Disconnect the hoses from the NIBP connector to enable proper zeroing. 3. Select Calibrate. If it is not available, perform the steps a and b. a. b. Select Protection OFF in the Calibration menu and push the Trim Knob. Menu selection Calibrate is now enabled. • Start calibration by pushing the Trim Knob. Messages ‘Zeroing’ and ‘Zeroed’ will be displayed in the NIBP message field. • Connect an external manometer with a pump to the module through the both tubes of the hose - both transducers B1 and B2 will be calibrated simultaneously. Pump up to a pressure of about 200 mmHg according to the manometer. Calibration is possible in the range of 150 to 250 mmHg. • Verify that both pressure values in the prompt field match the manometer reading. If not, adjust by turning the Trim Knob. When the values of the pressure bar and the manometer are equal, push the Trim Knob to confirm the calibration. The message ‘Calibrating’ and ‘Calibrated’ will be displayed on the NIBP digit field after a few seconds. • To set the protection on: Select Protection ON and push the Trim Knob. NOTE: If a long time not choose ON manually, this menu will automatically return to Protection ON. 5.2 Temperature calibration Temperature calibration should be performed: • • If the temperature test values differ for more than 0.1 °C. 1. Connect the TEMP accessories to the monitor, use P/N 884515-HEL Temperature calibration plugs 2. Enter STP Module service menu. After STP/TP board replacement. Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) Parameters - STP. 3. Enter Calibrations menu. 4. Choose Protection OFF in protect mode. 5. Select Calibrate T1/Calibrate T2. 6. Insert calibration plug (25 °C) into T1/T2 connector. 7. Push the Trim Knob. 8. Insert calibration plug (45 °C) into T1/T2 connector. 9. Push the Trim Knob. • Check Protection ON in protect mode. NOTE: If a long time not choose ON manually, this menu will automatically return to Protection ON. 3-25 Document no. 2081505-001 B40/B20 Patient Monitor 5.3 Invasive pressure calibration IBP calibration should be performed: • • • When the pressure transducer (probe) is replaced with a different type of transducer. When the measured value is not in the acceptable specification. After STP board replacement. NOTE: Before starting invasive pressure calibration, disconnect all patient cables and discharge the patient. There are two methods to complete the calibration as following: Using a transducer 1. Connect the IBP accessories to the monitor, use a pressure manometer with a pressure pump 2. Enter STP service menu. Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) Parameters - STP. 3. Enter Calibrations menu. 4. Choose Protection OFF in protect mode. 5. Connect a pressure transducer with a pressure manometer to the P1/P2 connector. Choose Calibrate P1 or Calibrate P2. Leave the transducer to room air pressure. 6. Push the Trim Knob to start zeroing. 7. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended pressure is 200 mmHg. 8. Set the pressure on the display to match the pressure reading on the manometer and push the Trim Knob. A tolerance of ±1 mmHg is allowed. 9. The message ‘Calibrated’ will be displayed on the display. • Check Protection ON in protect mode. NOTE: If a long time not choose ON manually, this menu will automatically return to Protection ON. 3-26 Document no. 2081505-001 Adjustments and calibrations Using a simulator 1. Connect the IBP accessories to the monitor and simulator. 2. Enter STP service menu. Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) Parameters - STP. 3. Enter Calibrations menu. 4. Choose Protection OFF in protect mode. 5. Choose Calibrate P1 or Calibrate P2. Set the P1 or P2 channel to 0 mmHg on the simulator. 6. Push the Trim Knob to start zeroing. 7. Set a pressure of 100 mmHg to 300 mmHg on the simulator. The recommended pressure is 200 mmHg. 8. Set the pressure on the display to match the pressure reading on the simulator and push the Trim Knob. A tolerance of ±1 mmHg is allowed. 9. The message ‘Calibrated’ will be displayed on the display. 10. Check Protection ON in protect mode. 3-27 Document no. 2081505-001 B40/B20 Patient Monitor 3-28 Document no. 2081505-001 4 Troubleshooting Introduction 1 Introduction The problems and solutions in this section represent only a few of the faults that you may encounter and are not intended to cover every possible problem that may occur. This chapter focuses on troubleshooting technical problems. For clinical and operation issues refer to the "User’s Guide" for troubleshooting. NOTE: For the Gas modules troubleshooting, please refer to “8. E-sCO, E-sCAiO and N-CAiO Modules“. NOTE: For the Entropy module troubleshooting, please refer to “9. E-Entropy Module“. If the problem remains, call technical support for service. To ensure accurate problem solving, please be prepared to provide the following information: • • • • Problem description and the troubleshooting done so far. Device information. Error messages displayed, if any. Other information, as requested. 1.1 Visual inspection Before beginning any detailed troubleshooting, complete a thorough visual inspection to be sure that: • • • There is no physical damage. The patient monitor is properly powered. No incompatible parts or devices used. Refer to “Supplies and accessories” for list of compatible devices. If loose parts or cable connections inside the monitor are suspected, disassemble the patient monitor to a level needed to perform an internal visual check. Check that: • • • all screws are tightened properly all cables are connected properly there are no loose objects inside the monitor NOTE: Perform the electrical safety test and the checkout procedure every time you have disassembled the patient monitor. 4-1 Document no. 2081505-001 B40/B20 Patient Monitor 1.2 General troubleshooting Disconnect the mains power cord and remove the batteries. No picture on screen Check the batteries charging levels by pressing the test buttons on the batteries. Connect the mains power cord Front panel green mains power LED lit? Replace fuses No Power board SW functioning Yes Insert only one battery with 2 or more capacity leds illuminating. Press the ON/OFF key. Front panel green battery LED lit? Yes Yes Disconnect the mains power cord and batteries. Reconnect and press the ON/OFF key. No No Mains power LED lit? Possible AC/DC unit failure. Replace the AC/DC unit. Power board failure Yes Mains power LED lit? Yes Wait for about 1 minute and press the NIBP auto ON/OFF key, NOTE: NIBP cuff must be connected NIBP pump starts pumping and connected SpO2 probe, Is SpO2 probe red LED lit? Possible key board or CPU failure. See Frame troubleshooting in this chapter. Display failure. See Frame troubleshooting in this chapter. Yes No Normal start-up sound and the alarm LEDs turn on and off? 4-2 Document no. 2081505-001 No No Main CPU board may be faulty. Introduction 1.3 Software troubleshooting chart Yes Start-up with GE logo image? Check the software CD whether can run in the PC normally and try to download the SW again. Turn off the monitor. Download the software from software CD onto the CPU board. NOTE: Refer to Software download instruction in CD. No OK? No Turn the power on. Wait for 120 seconds. No OK? Yes No Yes Start-up display appears? No Try with another software CD to download again. Please read the note below. Replace the CPU board and try again. Yes Yes Has the information regarding monitor software been updated on the Service View? Perform factory reset. NOTE: The software CD may be defective. 4-3 Document no. 2081505-001 B40/B20 Patient Monitor 2 Frame troubleshooting Problem Cause What to do Monitor is not starting. 1. The batteries are empty. 2. Fuses may be blown. 3. If power cord connected, AC/DC unit may be faulty. 4. If AC/DC unit is working, the power board may be faulty. 5. On/Stby key may be faulty. 1. Connect the power cord. 2. Replace fuses. 3. Replace the AC/DC power unit. 1. The connection between power board and CPU board may be faulty. 2. Faulty CPU board. 1. Check connection between power board and CPU board. 1. The display cables are loose. 1. Check the display connection board connectors. 2. Check inverter cable. Backlight Inverter may be faulty. Replace the Inverter board. Monitor is not starting. The monitor starts (alarm beep is heard), but the display remains black. 2. The backlights are not lit. 4. Replace the power board unit. 5. Replace the keyboard/membrane switch panel or the interconnection cable. 2. Replace the CPU board. Display and monitor operating 1. Loudspeaker connector or wires but no audible beep in start-up. loose or faulty. 1. Check loudspeaker connector and wires. Display is too dim. 1. Incorrect brightness adjustment. 2. Backlight faulty. 3. Backlight inverter faulty. 1. Adjust display brightness higher. 2. Check inverter cable. 3. Replace backlight. Stripes or white areas on screen. 1. Loose faulty display connection cable in CPU and display. 1. Check display connection cable in CPU and display. Module data disappears from the screen. ‘Module power supply overload’ message. Parameter module current (in module bus) too high. Detach and change parameter module. Module data disappears. 1. Module bus voltage or signals path broken. 1. Replace module interface board. 2. Replace the cable between module interface board and Power board. 3. Change the parameter module. ‘Replace Battery’ message on the screen. 1. Problem in communication between battery and power board. 2. Battery too old or defected. 3. power board may be faulty. 4. Problem in communication between power board and CPU 1. Replace battery. 'Frame temperature high' message. The temperature inside the frame is too high. Check monitor ventilation holes. ‘Battery temperature high’ message on the screen. Battery SMBus temperature is too high. Check monitor ventilation holes. Replace battery. 4-4 Document no. 2081505-001 2. Replace battery. 3. Replace the power board. Frame troubleshooting Problem Cause What to do Keyboard not working, but module communication is OK. Keyboard cables and connectors may be faulty. 1. Check the keyboard/membrane switch connection to the user interface board. 2. Check the interconnection cable between user interface board and battery board. Keyboard not working, and module communication not working. UPI section of the CPU board not functioning normally. Restart the monitor. Keyboard partly not working. 1. Check the interconnection cable 1. Keyboard/membrane switch between user interface board and faulty. battery board. 2. Keyboard cables and connectors 2. Check the keyboard/membrane may be faulty. switch connection to the user interface board. 3. Replace the keyboard/membrane switch. Replace the CPU Board. 4-5 Document no. 2081505-001 B40/B20 Patient Monitor 2.1 Battery indicators The messages, screen symbols and the LED indicators tell the user about the status of the batteries. The Green main LED will indicate that monitor is powered by Main. The Green battery LED will indicate that monitor is powered by battery. The Orange battery LED lit will indicate that the battery is in charging. The Orange battery LED flashing will indicate that the battery is defect. If the battery is fully charged and it’s main powered. the orange and Green battery LED will both dark. Table 4 Battery indicators Screen symbol A B Explanation Front panel battery LED indicators Monitor is battery powered. Batteries are fully charged and the size of the green bar indicates the charging level. Orange dark Green lit B Monitor is battery powered. Battery A is empty, battery B is ok. Orange dark Green lit B Monitor is battery powered. Battery A failure, battery B is ok. Orange flashing Green lit B B A NOTE: If both batteries fail, the green battery LED is dark. 4-6 Document no. 2081505-001 Monitor is mains powered. Battery A is being charged (white bar), battery B is already charged. Orange lit Green dark No battery backup. Monitor is mains powered. Orange dark Green dark Frame troubleshooting 2.2 NET section troubleshooting Problem Cause Monitor does not connect to Patch panel the network. Patch cable Monitor connects to the network, but disconnects unexpectedly (‘Network down: central station’ message on the monitor screen). What to do Patch cable not connected to HUB or to panel. Patch cable or connector defective. HUB not connected to power supply. HUB port closed due to physical layer problems. HUB port temporarily closed and reopened due to physical layer problems. HUBs not properly connected to each other. Monitor-Network cable Cable not properly connected to the wallplate or to the monitor. Cable or connector defective. Network cable (inside the monitor) defective Replace the Network cable - in FRU 2053489-004. External interface board defective Replace the External interface board - in FRU 2053489-004. Net section of the CPU board The NET section is defective. Replace the CPU board. NET section memory on the CPU board The SDRAM of the NET section is defective or uninitialized. The NET cannot be used. See network service page for details. ‘Check network connectors’ Monitor-Network cable message shows on the monitor screen Network printing fails Cable not properly connected to the wallplate or to the monitor. Cable or connector defective. Network cable (inside the monitor) defective Replace the Network cable - in FRU 2053489-004. External interface board defective Replace the External interface board - in FRU 2053489-004. Didn’t properly set up “virtual plug id“ Set up the virtual plug id in monitor. Print server is busy Network manager's print server is busy at the moment and cannot take more print jobs. Try again after 15 seconds. Print queue is full There are too many unprinted documents waiting in the print queue. Check the printer, as it is not operating properly. Printer is off-line Printer cable is loose, printer is out of paper, there is a paper jam or the printer is simply switched to off-line state. 4-7 Document no. 2081505-001 B40/B20 Patient Monitor Problem Cause What to do No vitial signs data send to EMR system No HL7 license Check about the HL7 license HL7 configuration error • Check HL7 configuration for vitial signs receiver IP and port. • Check HL7 Status view menu The vitial sign receiver system Check the Service Log, if have “AR” or “AE” have error or reject to receive. message from vitial sign receiver system, this is the receiver system’s issue. 4-8 Document no. 2081505-001 Hemo Troubleshooting 3 Hemo Troubleshooting 3.1 NIBP troubleshooting flowchart NIBP module not working Insert Hemo module to a good monitor Yes Yes Replace the flex module bus No Is it OK? Check the module interface board; the cable between module interface board and power board in the frame Remove the cover of Hemo module (connect the module to monitor, then power on) No Green LED on NIBP board lit? Replace NIBP board NIBP parameter display on the screen? No Yes Yes Start NIBP without hose Cuff loose message appears on screen? No Check if NIBP connector spring contact well with the button of cuff ID board Yes ENd NIBP function well? Leak test in Service Menu OK? No Check tubes connectors and manifold Pump check in Service Menu OK? No Check pump connector. No See error code explanation in service manual and fix it No NOTE: Please refer to 5.4. NIBP Module in Chapter 5 for Service Menu instruction. 4-9 Document no. 2081505-001 B40/B20 Patient Monitor 3.2 ECG troubleshooting flowchart ECG module not working No Refer to NIBP flowcharts No Replace the flex module cable. Or remove STP board (STP board may faulty) No End Yes Does NIBP module work well? Replace ECG board Yes ECG parameter display on screen? No ECG function well? Check ECG input connector, ECG input board and ECG input flex cable 4-10 Document no. 2081505-001 Hemo Troubleshooting 3.3 STP troubleshooting flowchart STP module not working No Refer to NIBP flowcharts No Select the right configuration No Replace the flex module cable. Or remove ECG board from the flex module cable. No Check STP input board, STP/TP input flex cable. No Check STP input board, STP/TP input flex cable. Check Masimo/Nellcor SpO2 board connect on the TP board well. Yes Does NIBP module work well? Go to service menu Yes Is STP parameter configuration correct? Replace STP/TP board Yes STP/TP parameter display on screen? Yes Temp/IBP function well? Yes SpO2 function well? End 4-11 Document no. 2081505-001 B40/B20 Patient Monitor 3.4 NIBP 3.4.1 NIBP toubleshooting Problem Cause What to do No NIBP value displayed NIBP not selected on screen. Check monitor setup. ‘Weak pulsation’ message Weak or unstable oscillation pulses due to: Check patient condition and retry. • artifacts • weak pulse pressure due to Use proper size of cuff. Check attachment. Check any leaks and retry. arrhythmias • improper cuff position or attachment • too few pulses detected • weak or unusual blood circulation • obese patient Call service ‘Error X’ message NIBP hardware error. X = error number. See 3.4.2. NIBP error code explanation for the description of the error message code. ‘Cuff loose’ message 1. Hose and/or cuff not connected. 1. Connect the hose and the cuff. 2. Hose and cuff connected. Reasons: 4-12 Document no. 2081505-001 - cuff loosely wrapped - Tighten the cuff. - leakage inside the shield, in the Patient connector panel or tubings connecting to the module - Check the tubings inside the shield and Patient connector panel, fix if necessary. - leakage in cuff or hose - Replace cuff/hose. - leakage inside module - Check internal tubing and fix if necessary. - pump does not work - Check pump connector; if OK, replace the NIBP Pump Unit. Hemo Troubleshooting Problem Cause What to do Air leakage 1. Hose or cuff leaking. Reasons: 1. Replace cuff - cuff damaged - Replace cuff. - cuff connector damaged - Replace cuff connector (if the fault is in hose connector). - O-ring damaged or missing - Replace O-ring. - hose double connector damaged - 2. Hose and cuff OK. Reasons: ‘Cuff occlusion’ message Replace NIBP cuff hose. 2. Connect or replace tube - leakage in the tubes connecting the patient connector panel and the module Check the tubes. - leakage inside the module - Replace the whole tubing. - tube disconnected or damaged - Fix connections. - manifold leaking - Replace the manifold. - tubes or valve(s) damaged - Replace tubes/valve(s). 1. Cuff and/or hose occluded. Reason: - cuff tube kinked - Straighten tube. - tubes inside the shield kinked - Straighten tubes. - tubes inside module kinked - Straighten tubes. - occlusion inside/outside module - Remove occlusion. 2. Cuff, hose, and tubes OK. Reason: - fault in pressure transducer - Replace the NIBP board. - fault in A/D converter - Replace the NIBP board. - faulty calibration - Check calibration. 4-13 Document no. 2081505-001 B40/B20 Patient Monitor 3.4.2 NIBP error code explanation Code Problem What to do 0 RAM test failure Change the NIBP board. 1 ROM checksum failure Change the NIBP board. 2 Pump on during idle or over current detected Check short circuits. Change the NIBP board. 3 Startup communication failure with safety CPU Change the NIBP board. 4 EEPROM protection is off Protect calibration by selecting Protection ON in the NIBP calibration menu. 5 EEPROM read/write error Change the NIBP board. 6 Valve stuck closed during cuff typing Try to remeasure. If the problem persists, recalibrate. If the problem still persists, change the NIPB board. 7 Could not save calibration data Reset the module and recalibrate. If this does not help, change the NIBP board. 8 PT2 higher than 150 for greater than 15 seconds Check short circuits. Change the NIBP board. while idle 9 Determination time too long Automatic recovery. 10 RTK 400Hz timer re-entry Change the NIBP board. 11 RTK 50Hz timer re-entry Change the NIBP board. 12 Not in use Not in use. 13 RTK overrun Change the NIBP board. 14 Too early AUTO START according to module check Reset the monitor. 15 Calibration data invalid on initialization or unit never calibrated Recalibrate. If this does not help, change the NIBP board. 16 Communication timeout between main and safety CPU Check short circuits. Change the NIBP board. 17 Safety CPU report communication timeout Check short circuits. Change the NIBP board. 18 Wrong message rate in communication between Check short circuits. Change the NIBP board. main and safety CPU ><+2% (480 msg/s) 4-14 Document no. 2081505-001 Hemo Troubleshooting 3.5 ECG Problem Cause What to do HR numerical display shows ‘---’ No heart rate available. If no ECG waveform, check LEADS OFF message and connect the leads. If ECG waveform exists, check heart rate source e.g. in the ECG Setup menu behind ECG key. Unacceptable ECG waveform No ECG trace Noise-message Poor electrode or poor electrode skin contact. Electrodes from different manufacturers are used. /Too much/little gel is used. Poor electrode condition. Electrodes are dried out. Improper site of electrodes. Check that electrodes are not placed over bones, active muscles, or layers of fat. Improper skin preparation. Remove body hair. Clean attachment site carefully with alcohol. Improper bandwidth filter. Check filter. Dirty ECG cable. Clean the cable. Faulty ECG cable Change the cable. Waveform not selected on screen. Press the Monitor Setup key and make adjustments. Module not plugged in correctly. Check the hemo module’s installation. High frequency or 50/60 Hz noise. Isolate noise source. 4-15 Document no. 2081505-001 B40/B20 Patient Monitor 3.6 Impedance respiration Problem Cause What to do No resp trace Waveform not selected on the screen Press the Monitor Setup key and make adjustments. Module not plugged in correctly Check the hemo module’s installation. Poor electrode or poor electrode skin contact Electrodes from different manufacturers are used. Too much/little gel is used. Poor electrode condition Electrodes are dried out. Improper site of electrodes Check that electrodes are not placed over bones, active muscles, or layers of fat. Improper skin preparation Remove body hair. Clean attachment site carefully with alcohol. Faulty/ dirty ECG cable. Change new cable. Respiration source is CO2 Check respiration source and change it to correct one. Unacceptable resp waveform APNEA message, and respiration waveform normal 4-16 Document no. 2081505-001 Hemo Troubleshooting 3.7 Pulse oximetry (SpO2) Problem Cause What to do Message ‘NO PROBE’ No sensor connected to the module SpO2 connector. Check sensor connections. Sensor faulty. Change the sensor. The type of sensor incorrect. Change the sensor. Flat cable connecting the SpO2 connector to the STP board loosen or broken. Check the Flat cable, replace if necessary. Unsuitable site. Try another site. Sensor faulty. Try another sensor. Sensor connection cable not connected to sensor. Connect the cable to sensor. Sensor is slippery. Wipe with 70% isopropyl alcohol and allow drying. Finger is too thin or thick. Try other fingers, or other sensor types. Poor perfusion. Try another place. Message ‘PROBE OFF’ though sensor properly attached to the patient Finger sensor falls off Weak signal artifacts Movement artifacts. Shivering. Message ‘CHECK PROBE’ Singal quality is too low to perform SpO2 measurements. Check probe condition or try another sensor. The type of sensor incorrect. Change the sensor. The monitor’s configuration for SpO2 is Check the monitor’s configuration. not correct. Message ‘FAULTY PROBE’ Sensor is faulty. Change the sensor. The type of sensor incorrect. Change the sensor. The monitor’s configuration for SpO2 is Check the monitor’s configuration. not correct. No SpO2 No waveform selected on screen. Check the selected SpO2 waveforms by pressing Monitor Setup key and selecting Screen Setup - Waveform Fields. Wrong configuration setting. Check the configuration settings from the STP/Calibrations menu (Monitor Setup - Install/Service - Service Parameters) 4-17 Document no. 2081505-001 B40/B20 Patient Monitor 3.8 Temperature Problem Cause What to do No temperature displayed Wrong type of probe. Use correct probe. Temperature out of measurable range. The range is between 10 and 45 °C. Temperature calibration not protected. Set the protection ON in the Service Menu. The monitor not configure TEMP. Check the monitor’s configuration. 3.9 Invasive blood pressure Problem Cause What to do Abnormally low pressure Transducer wrongly positioned. Check mid-heart level and reposition transducer. No pressure Defective transducer. Check transducer. ‘Not zeroed’ message No pressure module plugged in. Check the module. No waveform selected on screen. Check the selected pressure waveforms by pressing Monitor Setup key and selecting Screen Setup - Waveform Fields. Check that the pressure transducer is open to the patient. Wrong configuration setting Check the configuration setting from the STP/Calibrations menu (Monitor Setup - Install/Service - Service Parameters). Measurement on, channel not zeroed. Zero the channel. Invasive pressure calibration is not selectable Patient case is active Discharge a patient and make sure that the hemo module does not receive any signals from a simulator. Out of range < - 40 mmHg Measurement pressure is beyond the measurement range. Check the transducer level. Zero the channel. Out of range > 320 mmHg Measurement pressure is beyond the measurement range. Check the transducer level. Zero the channel. The patient may also have high pressure. Out of range Measured pressure is beyond the internal measurement range of the module. The waveform hits the top and the numeric display not shown. Check the transducer and its level. Zero the channel. 4-18 Document no. 2081505-001 5 Service Menu Introduction 1 Introduction The monitor has a Service Menu, which is a useful tool to examine monitor functions and troubleshoot in case a fault occurs. 1.1 Service Menu structure 5-1 Document no. 2081505-001 B40/B20 Patient Monitor 1.2 Service Menu NOTE: The Service Menu pictures are for reference only. Details on the menu page can vary depending on the software version and the module type in use. If a particular selection is not available in your system, the selection is shown grayed. 1. Press the Monitor Setup key. 2. Select Install/Service (password 16-4-34). 3. Select Service (password 26-23-8). Service Menu Sw version / Unit id SW Management Main Software --------------------------------LX/VSP_0.06 BootStrap software --------------------------B40_BOOTSTRAP_0.6 UBoot softeware -------------------------------B40_UBOOT_0.8 Linux kernel ----------------------------------B40_OS_0.9 File system ----------------------------------------B40_FS_0.6 UMBC software --------------------------------------B40_UMBC_0.6 CPU serial number: -----------------------------92114469 CPU test date: -- code: --- level: --2008-09-02 M1008748 06 PMC version -------------------------------------- Frame keyboard Parameters Set / Test Service Log Scroll Vers Record Vers Clear Password Previous Menu Scroll Vers Press Scroll Vers to scroll and view the detail version for monitor and modules on the extension window. Record Vers Press Record Vers to record the detail version for monitor and modules on the extension window. Clear Password The monitor will remember the service menu’s password you first enter. If you not restart the monitor, the next time you enter to service menu, any random password can get the access. Press Clear Password to clear the memory of password for service menu. When you next time enter to service menu, can use other level password for different service menu’s access. 5-2 Document no. 2081505-001 SW Management 2 SW Management The SW Management menu includes Software-specific service menus. Service Menu Sw version / Unit id SW Management ---------------------------- SW Download ---------------------------- Active Inactive SW Module SW Upgrade ---------------------------- Country Settings ---------------------------- License ------------------------------------------------------- Previous Menu ---------------------------number:----------------200 ---------------------------- 2.1 SW Download The SW Download submenu show IP Address and enable software download. IP Address: Show the IP Address of the monitor. Enable SW Download: Select YES can enable to download main software. SW Download IP Address 172.16.1.23 Enable SW Download Previous Menu Selecting this enables Software download 5-3 Document no. 2081505-001 B40/B20 Patient Monitor 2.2 Active Inactive SW The Active Inactive SW submenu allows you to active inactive software. Activation: Selection of activation can restart the monitor and change the main software from active software to incative software. NOTE: If there is no inactive software in the monitor, the selection is grayed. Activation Activation Previous Menu Soft Active Status Active SW B20/B40 Software: 0.19-0.6 Inactive SW B20/B40 Software: empty 2.3 Module SW Upgrade This submenu is used for module’s firmware upgrade. Current Version: to show the related parameter firmware current version. Upgrade Version: to show the related parameter firmware upgrade version. SW Upgrade: after you download the new file system file, use this menu to activate the new version of the related parameter firmware software. Upgrade SW Type: Select the parameter for firmware upgrade, the options are: NIBP, ECG, RECX. Module SW Upgrade Current Version B40_NIBP_1.04 Upgrade Version NA SW Upgrade Upgrade SW Type NIBP Previous Menu Return to previous menu. 5-4 Document no. 2081505-001 SW Management 2.4 Country Settings The Country Settings submenu can be refered to 3.1. Country Settings. 2.5 License The license submenu is used to set up monitor configuration for features. Please contact GE representative to order the upgrade kits, which include the related activation code for functions. Serial number: Show serial number for this monitor, this item can’t be selected. Product type: Show product type for this monitor: B40, B40i, B20, B20i. You can input the activation code to enable the following function: GESP: GE SpO2 NELL: Nellcor SpO2 MASI: Masimo SpO2 INVP: IBP TEMP: Temperature OCRG: OCRG NRES: Neonatal Respiration UNTY: Unity network NTS5: S5 network NHL7: HL7 network UTHL: Unity + HL7 network NONE: None network NOTE: When you finish to input the activation code, please manually perform a cold start, then all the settings will be activated. NOTE: If disable the IBP or TEMP function, the IBP and TEMP related settings in each mode will turn OFF, include the following settings: - Waveform Fields - Digit Fields - Graphical Trends pages - Snapshot fields - Record Waveforms The IBP and TEMP settings for Record Trends will change to Pleth. 5-5 Document no. 2081505-001 B40/B20 Patient Monitor 2.6 Enter/Exit Demo Mode The submenu allows you to enter/exit the demo mode. Under Demo Mode, the monitor displays the main vital signs values and waveforms. No need accessories, central station or any other peripheral equipment connect to the monitor while in Demo Mode. NOTE: After select this menu, the monitor will restart automatically. A “memory error” message will appear first time, please mannually restart the monitor again. NOTE: All the values and waveforms the Monitor displays are fictional. NOTE: The Demo Mode is only designed for the use of training and demo of operation. It is not intended for clinical use or patient monitoring and diagnosis. 5-6 Document no. 2081505-001 Frame 3 Frame The Frame menu includes frame-specific service menus. Service Menu Frame Sw version / Unit id ---------------------------- Country Settings Network ---------------------------- Power Supply ---------------------------- Previous Menu ------------------------------------------------------------------------------------------------------------number:----------------200 ---------------------------- 5-7 Document no. 2081505-001 B40/B20 Patient Monitor 3.1 Country Settings Languages submenu: See following 3.1.1. Languages National Reqs: Select software features that include national requirements. Power Frequency: Set the power frequency (50/60 Hz). This setting is used to filter out possible power frequency interference from parameter measurements. Time Format: Set the time format of the real-time clock (12 h or 24 h). Country Settings Languages National Reqs None Power Frequency 60Hz Time Format 24 h Previous Menu Load laguages and change current language. 3.1.1 Languages Language Select a language to be used during monitoring. NOTE: Service pages will always appear in English despite of this selection. NOTE: For language codes, see the table below. Country Settings Languages Language ENG.LNG Delete Language Delete Language ENG.LNG Delete a language file from the permanent memory of the monitor. Previous Menu POR.LNG Cancel Change language translation of screen texts from currently available language files. 5-8 Document no. 2081505-001 Frame Table 4 Language abbreviations in language file names Abbreviation Language CHI Chinese CZE Czech DAN Danish ENG English FRE French GER German HUN Hungarian ITA Italian NLB Dutch NOR Norwegian POL Polish POR Portuguese RUS Russian SPA Spanish SWE Swedish TUR Turkish FIN Finnish POB Portuguese (Brazil) 5-9 Document no. 2081505-001 B40/B20 Patient Monitor 3.2 Network Different network configuration (S/5 or MC) will have different menus. MC Network Network Config TCP/I P HL7 Previous Menu Network Status Unity: S5: HL7: Licensed Non-Licensed Non-Licensed Unit Name: Bed Name: RtClin Marking: NRtClin Marking: NRtNClin Marking: X D0107 26 8 0 Interfaces: Ethernet WLAN Not Connected Not Connected Current Interface: Eth_B40 Connection: Unity Network Free message bu ffers Current: 42 Minimum: 41 S/5 Network Dri Config Dri Comm TCP/IP Previous Menu Network Status Unity: S5: HL7: Non-Licensed Licensed Non-Licensed Virtual Plug Id: Transfer Mode: DRI Level: DRI/ETH 2009 Interfaces: Ethernet WLAN Not Connected Not Connected Current Interface: Eth_B40 Connection: Protocol stack restarted Normal: 0 Error: 0 Free message buffers Current: 256 Minimum: 255 5-10 Document no. 2081505-001 Frame 3.2.1 Network Config (MC) Unit Name: It is used for setting the unit name in the monitor. The default unit name is “X”. 7 characters at most. Bed Name: It is used for setting the bed name in the monitor. The default bed name is the last five characters of the MAC address, excluding the colon delimiter characters. 5 characters at most. RtClin Marking: It is used for tagging the following outgoing packets with a DSCP marking of 26 (011010): realtime clinical information (waveforms, parameters, alarms), Realtime network control information (time) NRtClin Marking: It is used for tagging the following outgoing packets with a DSCP marking of 8 (001000): non-realtime clinical decision support information (Admission, Histories, Full Disclosure, Printing) NRtNClinMark: It is used for tagging the following outgoing packets with a DSCP marking of 0 (000000): non-realtime, non-clinical decision support information (Service, InSite) MCS IP Address: It is used for setting the MCS’s IP address, It’s grey when network is HL7. Save Changes: It is gray if the changes have not been done to the menu information. NOTE: When save changes the monitor will require a restart except MCS IP address changed. 3.2.2 TCP/IP TCP/IP Status The TCP/IP Status view shows the general status of the TCP/IP set up. Data link: “NIC” for ehternet, DHCP Status: “Disabled“ or “Enabled“ depending on the DHCP status. The DHCP status related information displayed below. TCP/IP TCP/IP Confing Ping Previous Menu TCP/IP Status (Ethernet) Data Link: IP Address Subnet Mask Gateway IP DNS Server DHCP Server SLP NIC 172.16.1.54 255.255.0.0 172.16.254.254 Not available Not available Not available DHCP Status: Disabled Current State INIT REBOOT Execution State INIT REBOOT Lease Time 0:00:00 Time Left 0.00.00 Logins 0 5-11 Document no. 2081505-001 B40/B20 Patient Monitor TCP/IP config IP address: For setting the static IP address of monitor. Subnet Mask: For setting the static subnet mask of monitor. Default Gateway: For setting the static default gateway of monitor. Speed and Duplex: For setting speed and dupliex of monitor. Save Changes: Close menu and save the current menu information to permanent memory. Cancel Changes: Close menu without saving changes Ping IP Address: For configuring the destination IP address for the ping command. Ping: Enable the ping command by sending Internet Control Message Protocol (ICMP) echo request packets to the target host and wait for an ICMP response. The response should be show in the Ping Status view. 3.2.3 HL7 The HL7 setup information is on HL7 Status view. 5-12 Document no. 2081505-001 Frame HL7 config HL7 Receiver Ip: For setting HL7 receiver IP address. HL7 Receiver Port: For setting HL7 receiver port. HL7 Interval For setting HL7 interval value. HL7 HL7 Config HL7 Patient Class: For setting HL7 patient class. E : Emergency I : Inpatient O : Outpatient HL7 Receiver Ip HL7 Receiver Port 6000 HL7 Interval 1 min P : Preadmit HL7 Patient Class E R : Reocurring Patient Acknowledgment Enable B : Obstetrics Save Changes U : Unknown Previous Menu Acknowledgment: Enable/Disable acknowledgment check to HL7 server. Save Changes: Close menu and save the current menu information to permanent memory. Set HL7 Receiver Ip 5-13 Document no. 2081505-001 B40/B20 Patient Monitor 3.2.4 Dri Config (S/5) The DRI Level: For setting the monitor's network communication. The network communication needs match to the iCentral’s configuration. It should be selected to 2009. Network Dri Config Virtual Plug ID: For setting the same plug ID as in iCentral’s. Transfer Mode: For choosing the transfer mode: DRI/ETH or DRI/UDP Care Area: For setting the care area “OR” or “Non OR” for monitor. Debug server IP: For R&D purpose only. Dri Level 2009 Virtual Plug ID 7978 Transfer Mode DRI/ETH Care Area Non OR Debug Server IP Save Changes: Select "Save Changes" to take in use changes made in this page. Save Changes Previous Menu Push to select DRI Level. 3.2.5 Dri Comm (in S/5) Session layer Select Socket 1, 2, 3, 4 to view the status accordingly. Dest. name shows the name of the Central the monitor is connected to. Dest. id shows the Central Subnet. Dest. address shows the MAC address of the monitor network NIC in Central. Session layer Socket 1 Socket 2 Socket 3 Protocol specific situations: For R&D purpose only. 5-14 Document no. 2081505-001 Socket 4 Previous Menu Session Socket 1 Status Dest. name Dest. id Dest. address CENTRALMSF Central Subnet 1 0a:1a:00:00:03:bb Protocol special situations: Tx resent critical Rx non-critical duplicate Rx critical duplicate Rx non-critical with wrong msg num Watchdog traffic disconnect Missing ack disconnect Protocol errors 0 0 0 1 0 0 1 Frame Ethernet The Ethernet Status view shows the general status of the ethernet network communication. Driver: Ethernet chip name. Cable: Indicates if the ethernet cable is connected. EthernetAddr: Monitor’s ethernet address. Ethernet Previous Menu Speed: Indicates the current ethernet communication speed. Statistics In/Packets: Total number of received packets since last cold start. Statistics Out/Packets: Total number of transmitted packets since last cold start. Statistics In/Bytes: Total number of received bytes since last cold start. Statistics Out/Bytes: Total number of transmitted bytes since last cold start. The errors in Status view see following table. Table 1 Ethernet Status Driver Cable Ethernet Addt Speed (bits/s) DP83907 Connected 00:40:97:0b:01:fb 0 Statistics Packets Bytes In 2527 297776 Out 11327 9837268 Data errors CRC Frame 0 0 Transm. 0 BER 0 Hardware errors Intern. Missed 0 0 FIFO 0 Overrun 0 Ethernet errors Value Usage CRC errors (CRC) Number of received packets with incorrect checksum. Frame errors (Frame) Number of received packets with incorrect frame structure. Transmission errors (Transm.) Number or errors in packet Transmission. Notes Refers to physical layer problems. An erroneous packet has often both frame and CRC errors. BER errors (BER) Internal errors (Intern.) Internal error of the network board. Must always be 0. Missed packets (Missed) Number of received packets lost due to overload. Must always be 0. FIFO errors (FIFO) Internal error of the network board. Must always be 0. Overrun errors (Overrun) Practically the same as above. Must always be 0. 5-15 Document no. 2081505-001 B40/B20 Patient Monitor 3.3 Power supply 3.3.1 Power page The menu shows the voltages and temperature measured by the power board and CPU board. The values in the column under Mean are the mean values of last one second, the Min column shows the minimum mean value, and the Max column the maximum mean value of the voltages and temperature measured during the current power ON. The voltages and currents are measured by the power Board, except the four lowest under heading CPU Board A/D, which are measured by the CPU Board. +5V and +3.3V values come thus both from the power Board and CPU Board. System power and Module power are calculated by the power Board. Voltages ACDC AC/DC converter’s output voltage, used as Min Mean Max Power Supply Voltages monitor input voltage when the mains cord is 14.90 ACDC 14.90 14.90 connected. Range when present: 15.25…16.55V 0.00 EXTDC 0.00 0.00 Power page (Note: this includes the measurement inaccuracy). 12.00 Bat1 12.00 12.08 WPM Battery 12.24 Bat2 12.24 12.24 EXTDC Not used. 14.92 VSvs 14.92 14.92 Previous Menu Bat1 Battery A voltage measured at power Board. 0.00 VBoost 0.00 0.00 Range 9…12.6V for Li-ION battery 14.98 VMod 14.98 14.98 0.00 Bat2 Battery B voltage measured at power 1/0 Mod 0.00 0.00 5.04 +5V 5.04 5.04 Board. Range 9…12.6V for Li-ION battery 3.35 +3.3V 3.36 3.37 VSys System voltage at power board. This is the Currents Mod Current monitor input voltage measured at the power 0.31 0.39 0.33 Sys Current 0.87 1.34 1.23 board after input voltage selection. Range Bat Current 0.00 0.00 0.00 9…16.5V. Powers VBoost Not used. 12.58 13.87 13.38 System Power 4.35 4.40 4.84 Module Power VMod Supply voltage for modules. VBOOST is fed Temp C through a circuit breaker to VMOD. Range Power 38.77 38.77 38.77 13.8…16V. Therm Not CHG 0.00 0.00 0.00 Dummy CHG I/O Mod Not used. 0.00 0.00 0.00 CPU Board A/D +5V At power board. Range 4.8...5.3V. CPU temp ( C ) 0.00 0.00 0.00 +3.3V At power board. Range 3.15…3.45V. VSvs Out V 0.00 0.00 0.00 +3.3V Currents 0.00 0.00 0.00 +5V 0.00 0.00 0.00 Mod Current Current from module bus voltage VMOD. Depends on the module configuration. Sys Current Current from system voltage VSYS. Depends on the system configuration and battery charging. Bat current Current from or to the battery selected (discharge or charge). Measured at power Board. Depends on the system configuration and battery charging. Powers System Power Power from VSYS, calculated by power Board software. System power = VSYS * SYS Current Module Power Power from VMOD, calculated by power Board software. Module power = VMOD * Mod Current Temp Power1, Power2 Power supply unit temperature, measured at power Board. Therm Not CHG Not used. Dummy CHG Not used. CPU Board A/D CPU Temp Not used. VSYS_OUT Not used. +3.3VMeasured at CPU Board. Range 3.15…3.45V. +5V Measured at CPU Board. Range 4.75…5.25V. 5-16 Document no. 2081505-001 Frame 3.3.2 WPM Battery Batts This page contains information related to the batteries and power supplies. The power supply part is practically the same as in Power Page. Battery information includes also data measured by the smart batteries themselves and transmitted to the power Board via SMBus. SMBus is System Management Bus, a two-wire interface closely resembling IIC. SMBus is used for battery communication. Batts information section of the page has two columns: Batt1 for battery A data and Batt2 for battery B data. NOTE: Text ‘SMBus’ above Current (SMBus mA) line shows which battery is connected to the SMBus. Battery information Dev. Chem. Full Cap.(mAh) Device chemistry. The monitor supports only Li-ION batteries. Full charge capacity of the battery; capacity of the battery when it is fully charged. Rem. Cap. (mAh) Remaining battery capacity. Rel. St. of CHG (%) Relative state of charge of the battery. Expressed as a percentage of Rem. Cap. (mAh) / Full Cap.(mAh). Current (SMBus mA) WPM Battery Batts Smart Batt1 Smart Batt2 Clear Temp Maxs Previous Menu Battery current (discharge or charge) measured by the battery, transmitted via SMBus to power Board. Batts Information Dev . Chem. Full Cap. (mAh) Rem.Cap. (mAh) Rel .St. of CHG (%) Crrent (SMBUS mA) Voltage (V) Voltage (SMBUS mV) ICHG Low Batt Temp ( C ) Max Batt Temp (PMC) Max Batt Temp (SMBUS) Temps ( C ) Min Power 45. 11 CPU 0. 00 Voltages (V) ACDC 15. 00 ExtDC 0. 00 VSys 15. 00 Boost 0. 00 VMod 14. 53 I/O-VMod 0. 00 Currents (A) 0. 00 Batt 0. 30 Module 1. 65 Sys Powers (W) 12. 38 Sys 4. 68 Module Batt1 LION 0 0 0 SMBUS 0 11. 61 0 Batt2 LION 3345 3345 100 0 11. 69 12128 0. 00 32. 60 0. 00 Mean 45. 90 0. 00 31. 90 32. 20 32. 00 Max 45. 90 0. 00 15. 00 0. 00 15. 00 0. 00 14. 59 0. 00 15. 00 0. 00 15. 08 0. 00 14. 69 0. 00 0. 00 0. 32 1. 16 0. 00 0. 32 1. 16 13. 52 4. 79 13. 52 4. 84 Voltage (V) Battery voltage measured at the power Board Voltage (SMBus mV) Battery voltage measured by the battery. ICHG Charging power level for charger hardware, this bit can have values high or low. High is the normal setting, low is used when the power Board software determines to limit the total power consumption of the monitor by limiting the charging power (i.e. due to high temperature). Batt Temp (°C) Battery temperature. This is real time data for the battery connected to SMBus. Max Batt Temp (PMC) This is subject to change. Max Batt Temp (SMBUS)Maximum battery SMBus temperature from entering the service pages. Max values are updated in real time for the battery connected to the SMBus. Other measurements See explanation in the previous power pages. 5-17 Document no. 2081505-001 B40/B20 Patient Monitor Smart Batt1 Battx information from SMBus (this information is received from the battery via SMBus) WPM Battery Temperature: Battery temperature Batts Voltage: Battery voltage Smart Batt1 Current: Battery current (discharge or charge) Smart Batt2 Avg. Current: Rolling average of the battery current Clear Temp Maxs Previous Menu Rel. State of Charge: Relative state of charge of the battery. Expressed as a percentage of Rem. Cap. (mAh) / Full Cap.(mAh). Abs. State of Charge: Absolute state of charge. Expressed as a percentage of Rem. Cap. (mAh)/ Design Capacity (mAh). Batt1 information from SMBus Temperature Voltage Current Avg. Current Rel. State of Charge Abs. State of Charge Remaining Capacity Full Charge Capacity Cvcle Count Design Capacity Design Voltage 30.75 C 11903 mV 0 mA 0 mA 97 % 72 % 2532 mAh 2604 mAh 33 3520 mAh 11100 mV Manufacture Date(DD:MM:YY) 20/12/24 Manufacture Name Device Name Device Chemistry Therm. Status from charger NPC A07A90 SM201-6 LION Remaining Capacity: Remaining battery capacity (mAh). Full Charge Capacity: Capacity of the battery when it is fully charged. Cycle Count: Number of cycles the battery has experienced. A cycle is an amount of discharge approximately equal to the value of Design Capacity. The exact value of cycle count threshold is stored in the battery permanent memory. Design Capacity: Theoretical capacity of a new battery. Design Voltage: Theoretical value for nominal voltage of a new battery. Manufacture Date (DD:MM:YY) : The date the battery pack was manufactured. Manufacturer Name: Acronym of the battery pack manufacturer name. Device Name: Battery pack model name. Device Chemistry: Battery chemistry of the cells used. Therm. Status from charger: Status of the battery thermistor or code resistor read by Smart Battery Charger IC. The thermistor or code resistor is always connected to the charger whenever the corresponding battery is connected to the charger and SMBus. 5-18 Document no. 2081505-001 Frame Smart Batt2 If Smart Battx is selected for a battery NOT connected to the SMBus, the menu contains the following: Battx information from Memory and PMC (This information comes from the power Board memory or is measured by the power Board. SMBus data in this menu is not real time, because this battery is not connected to the SMBus). NOTE: This page may not contain information if SMBus has been connected only to the other battery. Page can be updated by battery disconnection and reconnection, if desired. Rel. State Charge: Relative state of charge of the battery. Expressed as a percentage of Rem. Cap. (mAh) / Full Cap.(mAh). WPM Battery Remaining Capacity: Remaining battery capacity (mAh). Smart Batt1 Full Charge Capacity: Capacity of the battery when it is fully charged. Design Capacity: Theoretical capacity of a new battery. Batts Smart Batt2 Batt2 information from Memory and PMC Rel. State of Charge Remaining Capacity Full Charge Capacity Design Capacity Design Voltage Manufacture Date (DD:MM:YY) 0% 0 mAh 0 mAh 0 mAh 0 mV 00/00/00 Clear Temp Maxs Previous Menu Design Voltage: Theoretical value for nominal voltage of a new battery. Manufacturer Date (DD:MM:YY): The date the battery pack was manufactured Manufacturer Name Device Name Device Chemistry DC/DC Board A/D Batt1 Voltage Batt2 Voltage Batt Current Batt Not CHG Temperature LION 12.00 V 12.34 V -0.01 A 32.24 C Manufacturer Name: Acronym of the battery pack manufacturer name. Device Name: Battery pack model name. Device Chemistry: Battery chemistry of the cells used. Batt1 Voltage: Battery A voltage measured at the power Board. Batt2 Voltage: Battery B voltage measured at the power Board. Batt Current: Battery current (discharge or charge) for the battery connected to SMBus. Measured at the power Board. Batt Not CHG Temperature: This is subject to change. Clear Temp Maxs This command is useful only when the WPM Battery Batts view is selected. The Clear Temp Maxs command clears the maximum values of Batt Temp (SMBus) and Batt Temp (PMC). Note: Power temp and CPU temp maxs are not cleared. 5-19 Document no. 2081505-001 B40/B20 Patient Monitor 4 Keyboard The service menu for testing the command board functions. Red Led is for testing the red alarm light on the monitor. When the text is highlighted, the red alarm light can be turned on and off by pressing the Trim Knob. Yellow Led is for testing the yellow alarm light on the monitor. When the text is highlighted, the yellow alarm light can be turned on and off by pressing the Trim Knob. Cyan Led is for testing the cyan alarm light on the monitor. When the text is highlighted, the cyan alarm light can be turned on and off by pressing the Trim Knob. Dummy Press is for testing the Trim Knob. When the text is highlighted, pressing the Trim Knob creates a sound from the loudspeaker and the corresponding number on the service data field increases. Keyboard Service Data Red Led Message count 0 Leds red OFF yellow OFF cyan OFF Yellow Led Direct action keys Cyan Led Dummy Press Admit/ Pt. Data SpO2 Dischar & Trends Keyboard Log NIBP Previous Menu Invasive Normal Pressures Screen Monitor Setup Print/ Record Airway Gas Others Silence Alarms Alarms Setup Zero All Auto On/Off Start Cancel Start Stop Control wheel Press 0 Clokwise 0 Counterclokwise Service Data Message Count counts the number of messages that are sent out to the main CPU board. Leds upper and lower indicate the states of the alarm lights on the monitor. Direct action keys texts are indications to the command board membrane keys. When a key on the command board is pressed, the corresponding text in the menu changes its color. Control wheel, Press counts the Trim Knob pressings. Control wheel, Clockwise and Counter clockwise indicate the Trim Knob turnings. 5-20 Document no. 2081505-001 ECG 0 Keyboard 4.1 Keyboard Log All the keyboard presses and the commands given by the Trim Knob are recorded in the Keyboard Log. The length of the log is 80 events. The log is FIFO type. Scroll Log: enables to scroll the keyboard log. Scroll Stat: enables to scroll the keyboard presses events. Record Log: to record keyboard log. Record Stat: to record keyboard stat. Reset Log: to reset the keyboard log. Keyboard Keyboard Log Scroll Log Scroll Stat Keyboard : Keyboard Log 2004- Jan- 29 05:40:54 Record Log Service Menu : Keyboard 2004- Jan- 29 05:40:40 : Previous Menu 2004- Jan- 29 05:40:39 Record Stat Display Reset Log Previous Menu Service menu Frame Network Network Config TCP/IP TCP/IP Config : Display 2004- Jan- 29 05:40:11 : Previous Menu 2004- Jan- 29 05:40:09 : Previous Menu 2004- Jan- 29 05:40:05 : Previous Menu 2004- Jan- 29 05:40:00 : Previous Menu 2004- Jan- 29 05:39:57 : Previous Menu 2004- Jan- 29 05:39:40 5-21 Document no. 2081505-001 B40/B20 Patient Monitor 5 Parameters NOTE: Parameter values in Service Data fields are only for reference in this section. 5-22 Document no. 2081505-001 Parameters 5.1 Gas Unit 5.1.1 General The monitor starts counting these items at power up and resets to zero at power off. The values may also be reset when a module is attached to the monitor frame and be set to 32769 or continuous counting may be started when the module is removed from the monitor frame. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) or value 32769 indicates either a serial communication failure or a module not in place. Also failures in other modules may cause these numbers to rise or be set to 32769. Module configuration shows which measurement options are available, i.e. are detected by the module. Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor's inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to monitor has broken down. General Previous Menu Service Data Module configuration MiniC CO2 O2 N20 AA id p&v GasExch 1 0 0 0 0 0 0 0 0 = not available 1 = available Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. Timeouts -12867 Bad checksums 0 Bad c-s by moc 0 5-23 Document no. 2081505-001 B40/B20 Patient Monitor 5.1.2 Gases Noise Meas activates the noise measurement. Sample gain adj adjusts the sampling pump gain, i.e. for adjusting the sample flow measurement. Pump ctrl A manual control for the sampling pump. Zero valve ctrl A manual control for the zero valve. Gases Noise Meas Sample gain adj Pump ctrl Zero valve ctrl Previous Menu Serivce Data OFF noise-% mv Gain O2 --------CO2 0.00 300.0 2826 1.010 N20 --------AA1 A AA2 ----- B --ID --C --ID unrel. --D --E --MAC Sample Flow 148.7 Zero 0.0ml/min Gain 1.000 Ambient 1013 Amb- Work 49.6mbar % OFF Fall time CO2 CO2-02 Delay --02 0ms ---ms Pump ON 0.97% 1881mV Lamp ON 56.23% 75mA Fan ON Zero Valve MEAS Occl Valve MEAS Temp TPX 45.6 CPU 0.0 Time after power on OM 54min 0.0 Service Data O2, CO2, N2O, AA % shows real-time concentrations. noise-% is standard deviation of concentration. O2, CO2, N2O, AA channels A-E mV: signal is scaled to mV. Gain User gain. It is scaled as (User gain)/(Factory gain). ID Identified agent. ID unrel. The shown value tells how unreliable the identification is. With pure agent the value is normally <50. MAC MAC value Sample Flow calculated from differential pressure and adjusted by the module. Zero value as measured during initialization when the pump is off. Gain: sample flow measurement can be calibrated by adjusting the gain. Ambient Ambient pressure is measured every 30 min. Amb-Work Ambient pressure - sampling system internal pressure. Fall time CO2 and O2 in ms. For N2O and AA same as CO2. CO2-O2 Delay In ms. No delay between CO2, N2O, and agents. Pump Can be toggled ON/OFF. PWM output 0-100% is shown. Pump voltage is also shown. Lamp The state, PWM control, and current of the lamp are shown. Fan The state of the fan is shown. Zero and Occl Valve Can be toggled between measurement state (MEAS) and zeroing/occlusion states (ZERO/OCCL). Temp Temperatures measured by the module from TPX, CPU, and OM. Time after power on In minutes after power on. 5-24 Document no. 2081505-001 Parameters 5.2 ECG Module Power Freq is used to changing the power frequency for ECG module, it can display in Service Data’s view. To press this menu, change 50 Hz to 60 Hz, or 60 Hz to 50 Hz. Service Data Power Freq shows the mains frequency selected: 50 Hz / 60 Hz ECG Module Serivce Data Power Freq Cable type Electrode 50 Hz Quick Zero 5 lead Artifact RA LA LL V1 ON ON ON OFF V2 V3 V4 V5 0FF OFF OFF ON 0 0 RL ON V6 OFF Pacer vount 0 R count 4625 Electrode shows ON when each of these electrodes is connected. Resp: Avail Zeroing 1 0 Value MeasOff 20 0 Pacer count is a running number of detected pacemaker spikes. Arrhythmia HR 80 QRS count Det leads1 Data -73 0 HR25 4607 6 -12 On/ Off 80 HR75 PVC nroleads Noise M/A 1 80 0 2 0 RAM ROM EEPROM MissedPkgs OK OK OK 4 Cable type shows the leadwire set connected: 3 lead / 5 lead ECG Setup Quick Zero shows 1 when the ECG signal is beyond scale, and therefore, is quickly returned to the optimal range with fast signal processing methods. Previous Menu Power Freq Artifact shows 1 when artifacts are detected. R count is a running number of detected R waves. Resp: Avail shows if a module with impedance measurement is plugged (1) or not plugged (0) into the monitor. Value shows the measured respiration rate value for impedance respiration. Timeouts 6 Bad Checksum 0 Bad cs-by mo 0 Leads Off 0 Zeroing indicates the zeroing status of the respiration measurement: 1 = zeroing, 0 = not zeroing. MeasOff shows 1 if the respiration measurement is set to OFF, and 0 if the the respiration measurement is set to ON. Arrhythmia shows an active arrhythmia alarm (VFIB, VTAC, ASY). On/Off shows 1 if arrhythmia analysis is active and 0 if it is not active. HR, HR25 and HR75 for R&D purpose only QRS count shows the running number of detected QRS complexes. PVC shows the PVCs detected per minute Det leads shows the leads that are being used for detecting beats and ventricular fibrillation. The selection of user leads (ECG1, ECG2, ECG3) on the monitor affects the leads used for detection. The first lead used for detection is lead either I or II. The algorithm uses the lead appearing first in user leads. The second lead used for detection is one of the precordial leads (V1 - V6): -1 = invalid (not used), 0 = lead I, 1 = II or III; 2 = V1, 3 = V2, 4 = V3, 5 = V4, 6 =V5, 7 = V6 nroleads shows the number of leads that are beign used for detecting beats and ventricular fibrillation. Possible values include: 0 , 1 and 2 (ref. Det leads) analysis. Data for R&D purpose only Noise M/A for R&D purpose only Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor’s inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to the monitor has broken down. 5-25 Document no. 2081505-001 B40/B20 Patient Monitor Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The values may also be reset when a module is attached to the monitor frame and be set to 32769 or continuous counting may be started when the module is removed from the monitor frame. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) or value 32769 indicates either a serial communication failure or a module not in place. Also failures in other modules may cause these numbers to rise or be set to 32769. RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error). LeadsOff indicates whether the monitor can measure ECG even if one or more leadwires are off: 1 = measurement is not possible, 0 = measurement can be done. MissedPkgs indicates the number of packages missed. 5.2.1 ECG Setup Filter filters the ECG signal high frequency noise and slow respiratory artifacts: - - - ECG Module Monit (monitor) filter is used in routine monitoring. It effectively filters the artifacts caused by the electrosurgery unit and respiration. Diagn (diagnostic) filter is used if more accurate information of the waveform is needed (e.g. of P-wave or AV block). The diagnostic filter is more susceptible both to high frequencies and baseline wander than the monitor filter. STfilt (ST filter) permits more accurate information of ST segment. It filters the high frequency artifacts caused by the electrosurgery unit, but catches the slow changes in ST segment. The ST filter is more susceptible to baseline wander than the monitor filter. Pacemaker selects how to display the pacing pulse of cardiac pacemaker. The selections are Show, Hide and Sensit: - Hide: the pacing pulse is filtered away from ECG data. - Show: the pacer pulse is filtered away from ECG data but the pulse is displayed as a constant height marker. - Sensit: uses a more sensitive pacemaker detection. Pacemaker spike is displayed on ECG. 5-26 Document no. 2081505-001 Service Data ECG Setup ! Filter STfilt Pacemaker Hide Monit STfilt Previous Menu Diagn Quick Zero 0 Artifact 0 LL V RL ON OFF ON V4 V5 V6 OFF OFF OFF Rcount 0 Value MeasOff On/Off --1 0 HR75 --- PVC 0 nroleads 0 Noise M/A 0 0 Use ST filter for optimal ST analysis. RAM ROM EEPROM MissedPkgs OK OK OK 0 Parameters 5.3 STP Module (for GE SpO2) Record Data prints out the shown service data and board information (id, serial number and sw id) onto the recorder. Temp Test activates the automatic temperature test for the temperature channels T1 and T2. The result from the test is shown in the service data field. NOTE: The Temp Test needs to be selected twice before the test starts. Service Data field Gain is a coefficient to compensate gain error. Usually the values for P1 and P2 are between 17000 and 25000 and for T1 and T2 between 13000 and 14300. Zero indicates the offset compensation value of each parameter in the A/D converter. Typically the values for P1 and P2 are within ±1000 and for T1 and T2 between -150 and +300. Calibrate if zero and/or gain value is outside the ranges. STP Module Calibrations Record Data Temp test Previous Menu Cable shows ON when a corresponding cable is connected to the front panel Probe shows ON when a corresponding probe is connected to the cable. Value shows the measured numeric values simultaneously. Pressure values are real time values and shown in mmHg. Temperature values are shown in degrees Celsius. Serivce Data Gain Zero Cable Probe Value P1 P2 T1 11161 11163 -7562 -10295 -10295 9674 ON ON ON ON ON ON 76./4 /.94 3/.09 Buttons OFF OFF SpO2 Modpr Hr Cable Probe OK ------ON ON Ired Int. Red Int. DC gain IDC RDC AC gain Pre gain OFF Temp error Temp test T2 -7569 9695 ON ON 3/.09 71 70 5 14115 11193 0 0 OFF OFF Protect key ON Protect mode ON Configuration STP Timeouts Bad checksums Bad c-s by mode SpO2 shows the measured beat-to-beat SpO2 value. Modpr is a modulation % that indicates the AC/DC ratio in the measured signal. 2 0 0 RAM ROM EEPROM OK OK OK Hr is a pulse rate calculated from every beat. Cable and Probe can be either OFF or ON, and these indicate the state PROBE OFF. Under them there is a message field for SpO2. It can be OK, PULSE SEARCH, CBL OFF, PROBE OFF, NO PULSE, ARTEFACT, POOR SIGNAL, or CHECK PROBE. Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (Ired Int.) is in the range of 40 to 255 and red intensity (Red Int.) is in the range of 40 to 255. DC gain shows the gain of DC signal adjusted by the module. IDC is the value of infrared signal. RDC is the dc value of red signal. AC gain is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high ac gain and 0 means low gain. Pre gain is a preamplifier gain for infrared and red signals. Pre gain values can be 1 or 0. Value 1 means normal operation. Value 0 means that signal levels are very low and extra gain is taken into use. 5-27 Document no. 2081505-001 B40/B20 Patient Monitor Temp error shows the status of the temperature test. No errors found show the status (OFF) and errors found (ON). Protect key and Protect mode show normally ON but can be turned to OFF for the temperature calibration in Calibration Menu. Configuration shows the chosen module configuration: T, P, TP, ST, or STP. Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor’s inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to the monitor has broken down. Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either a serial communication failure, or a module not in place. Also other modules can cause communication errors that cause these numbers rise. RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error). 5-28 Document no. 2081505-001 Parameters 5.3.1 Calibrations Protection: Protection for the configuration and temperature calibrations can be set ON and OFF. STP Module Service Data Calibrations Calibrate T1 / Calibrate T2: The functions are for calibrating the temperature channels T1 and T2. Protection Calibrate P1/ Calibrate P2: The functions are for calibrating the invasive blood pressure channels P1 and P2. ON Calibrate T1 Calibrate T2 Calibrate P1 Calibrate P2 Previous Menu T2 T1 5185 15196 33 34 ON ON ON ON : 6./ 4 3/ .05 OFF int. 220 int. 220 ain 110 2047 2047 OFFin 0 gain 1 OFF Calibrate transducer with manometer. push Comwheel to start zeroing. 2 0 0 OFF OK RAM OK ROM EEPROM OK How to calibrate T1/ T2 The calibrations are possible only when the protection is set OFF. The temperature calibration requires accurate test plugs of value 25 °C and 45 °C. 1. Select Calibrate T1/Calibrate T2 2. Insert the test plug 25 °C into the T1/T2 connector 3. Press the Trim Knob 4. Insert the test plug 45 °C into the T1/T2 connector 5. Press the Trim Knob How to calibrate P1/ P2 NOTE: Before starting the pressure calibration, disconnect all patient cables and discharge the patient. The calibrations require a pressure transducer (with appropriate cable) and a pressure manometer. 1. Connect the pressure transducer with the pressure manometer to the P1/P2 connector. Select Calibrate P1/Calibrate P2. Leave the transducer to room air pressure. 2. Press the Trim Knob to start zeroing. 3. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended pressure is 200 mmHg. 4. Set the pressure on the display to match the pressure reading on the manometer and press the Trim Knob. 5-29 Document no. 2081505-001 B40/B20 Patient Monitor 5.4 NIBP Module Service Data Pressure shows the measured pressure multiplied by 10. This value is automatically zero-drift compensated. Zero shows the difference between the zeroing value in the permanent memory (stored when the module is calibrated) and the current automatic zero-drift compensation multiplied by 10. The value can change between +20 and -20 mmHg. If the zero drift exceeds ± 10 mmHg, the module should be recalibrated. AD0 to AD7 show the values of each eight channels of the A/D converter. ST1: Master status, indicate measurement, calibration, zeroing whether ongoing ST2: Button status, Not used yet. ST3: Hardware status, indicate the power supply status, ADC error status. ST4: Measurement result status, indicate whether measurement is ready, whether cuff ID is updated NIBP Module Calibrations Safety Valve Pnematics Previous Menu Service Data Pressure Zero B1 000000 -00010 B2 000000 000000 St1 St2 St3 St4 0000 0000 0400 0000 AD0 -17 6 AD1 -1 AD2 AD3 1502 2 AD4 AD5 -1644 5 AD6 AD7 -1505 ON OFF OFF Offset Gain Timeouts Bad checksums Bad c-s by mod B1 000103 000041 B2 000096 000041 2 0 0 OK RAM OK ROM EEPROM OK Offset show the offset values for B1 and B2 ( pressure sensor calibration factors). Gain show the gain values for B1 and B2 ( pressure sensor calibration factors). Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor’s inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to the monitor has broken down. Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either a serial communication failure, or a module not in place. Also other modules can cause communication errors that cause these numbers rise. RAM indicates the state of the RAM memory. ROM indicates whether the checksum in the EPROM is in accordance with the one the software has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error). 5-30 Document no. 2081505-001 Parameters 5.4.1 NIBP Calibration Active Leak Test: Wrap an adult cuff around a pipe and connect the cuff to the module. Select the active leak test (ON). The module automatically pumps a pressure of 280 mmHg into the cuff. Wait for several seconds until the pressure stabilizes. Then check that the pressure reading does not drop more than 6 mmHg per minute. If it does, leaking point(s) should be detected and fixed. Cancel the test by selecting the Active leak test OFF. NIBP Module Service Data Calibration Active Leak Test OFF Calibration Check OFF Protection OFF B1 00 10 AD0 -17 6 AD1 -1 AD2 AD3 1502 2 AD4 AD5 -1643 5 AD6 AD7 -1505 Calibrate Previous Menu Calibration Check: After the calibration check is selected (ON), the module zeroes the pressure transducers at the beginning of the calibration check. Do not pump pressure until the text ‘Zeroed’ appears in the NIBP digit field or the zeroing will fail. After the zeroing is done, manually pump pressure into the module and make sure that the same pressure values are shown both on the display and on the manometer. Pressure of both pressure channels B1 and B2 are shown. The pressure values are automatically zero-compensated, so the readings of B1 and B2 should be the same as the manometer readings. B2 000000 000000 2 0 0 OK RAM OK ROM EEPROM OK Protection: Software calibration protection (ON/OFF). Select OFF when calibrating. How to calibrate NOTE: Perform NIBP Calibration Check first to evaluate if calibration is needed or not. NOTE: Both transducers B1 and B2 will be calibrated simultaneously. NOTE: The module must be in the frame during the whole procedure. NOTE: Calibration selection is available only when protection is OFF. 1. Change the protection setting from ON to OFF to enable the Calibrate selection - the color of the Calibrate selection changes from grey to white. 2. Zeroing: − Disconnect the NIBP hose from the module connector. − Select Calibrate and push the Trim Knob. NOTE: Messages ‘Zeroing’ and ‘Zeroed’ is shown in the NIBP message field and next to the Calibrate selection momentarily. After this, a pressure bar will appear beside the menu. 3. Calibration: − Connect the NIBP hose to the module connector. − Connect an external manometer with a pump to both tubes of the hose. − Pump about 200 mmHg pressure. − Verify that both pressure values, B1 and B2, shown in the prompt field of the calibration menu match the manometer reading. If not, adjust the by turning the Trim KnobTrim Knob. − Press Trim Knob to complete the calibration. 5-31 Document no. 2081505-001 B40/B20 Patient Monitor NOTE: Messages ‘Calibrating’ and ‘Calibrated’ are shown in the NIBP message field and next to the Calibrate selection. NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the Trim Knob, even in cases where the value would be correct. For example, change the value one step higher and then back one step lower. The ‘Calibrated’ text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module. 4. Change the Protection setting from OFF to ON to disable Calibrate selection - the color of the Calibrate selection changes from white to grey. NOTE: If a long time not choose ON manually, this menu will automatically return to Protection ON. 5.4.2 NIBP Safety Valve Start test is for starting and Stop test for stopping the Safety Valve test. Safety Valve Data: For information on general items Pressure, Zero, St1 to St4, AD0 to AD7 as well as Timeouts etc., see service data descriptions in section 4.4 NIBP Module. Safety Valve ADULT Start Test Previous Menu Safety Valve Data Pressure Zero B1 000000 000000 B2 000000 000000 St1 St2 St3 St4 0000 0000 0400 0000 AD0 -16 6 AD1 -1 AD2 AD3 1502 2 AD4 AD5 -1643 4 AD6 AD7 -1505 Max. press and 2 s after stop show the measured values at Safety Valve test. Max press 2 s after stop Timeouts Bad checksums Bad c-s by mod B1 0 0 B2 0 0 2 0 0 OK RAM OK ROM EEPROM OK How to do safety valve test 1. Disconnect the pressure manometer from the NIBP cuff connector. Connect the NIBP hose and cuff to the NIBP cuff connector. 2. Connect a standard adult cuff around some round object, for example a calibration gas bottle. 3. Perform a NIBP test by pressing NIBP Start/Cancel button (Waiting until the test ended). 4. Select Start Test. Start the adult safety valve test by pressing the Trim Knob. Wait until the pump stops and the pressure is deflated. 5. Check the pressure values ‘Max press’ for both transducers. All the values should be within 300 - 330 mmHg. 6. Connect a neonatal hose and cuff around another round object, for example a calibration gas bottle. 7. Perform a NIBP test by pressing NIBP Start/Cancel button (Waiting until the test ended). 8. Select Start Test. Start the neonatal safety valve test by pressing the Trim Knob. Wait until the pump stops and the pressure is deflated. 5-32 Document no. 2081505-001 Parameters 9. Check the pressure values ‘Max press’ for both transducers. All the values should be within 150 - 165 mmHg. 10. Return to the normal monitoring mode by pressing Normal Screen. 5.4.3 NIBP Pneumatics Start Pump/Stop Pump: A manual control for the pump. The selection changes to Stop Pump when the pump turns on. Open Safe. Valve/Close Safe. Valve: A manual control for the Safe valve. The selection changes to Close Safe Valve when the valve is opened. Open Defl. Valve/Close Defl. Valve: A manual control for the deflation valve. The selection changes to Close Defl. Valve when the valve is opened. Pneumatics Start Pump Pneumatics Data Pressure Zero B1 000000 -00010 St1 St2 St3 St4 0000 0000 0400 0000 B2 000000 000000 Open Safe. Valve Open Defl. Valve Reset Clock Previous Menu Reset Clock: Set up the Interval time 20 mmHg -> 185 mmHg to 0 s Pump Safety Valve CLOSED OFF Pneumatics Data field For information on general items Pressure, Zero, St1 to St4, AD0 to AD7 as well as Timeouts etc., see section "NIBP Module." AD0 -17 6 AD1 -1 AD2 AD3 1502 2 AD4 AD5 -1643 4 AD6 AD7 -1505 Deflate Valve CLOSED Interval 20 mmHg -> 185 mmHg Timeouts Bad checksums Bad c-s by mod 2 0 0 0s OK RAM OK ROM EEPROM OK Pump, Safety Valve, and Deflate Valve show their states. How to check Interval 20 mmHg -> 185 mmHg Select Start pump at different combinations of the valves open/closed and push the Trim Knob. The module counts the time it takes for the pressure to go up from 20 mmHg to 185 mmHg and displays it. When all the valves are closed, the pump should be able to pump the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not stop without selecting Stop Pump by pushing the Trim Knob. NOTE: To redo the test, must go back to the previous menu. 5-33 Document no. 2081505-001 B40/B20 Patient Monitor 5.5 SpO2 (for Masimo/Nellcor SpO2) SpO2 Data PR: Measured pulse rate value SpO2%: Measured SpO2 value NoProbe: If there is a probe connected SpO2 Previous Menu PulseSearch: If the pulse search is being done. SpO2 Data PR SpO2% --------- NoProbe PulseSearch CheckProbe 1 0 0 QUART Status POX Status I/O Status POX Error 0000 0000 9E08 0000 CheckProbe: If there is check probe error. QUART Status: Show quart status. POX Status: Show POX measurement status. I/O Status: Show IO status. POX Error: Show POX error status. For information on Timeouts etc., see section "NIBP Module." Return to Previous menu. Timeouts Bad checksums Bad c-s by mod 2 0 0 ROM OK 5.6 Entropy When the Entropy Service Menu is open, the function of Entropy module keys will be disabled. Check Sensor: to replace module keys function to check the entropy sensor. Module: to show the Module status in the right window. Sensor: to show the Sensor status in the right window. 5-34 Document no. 2081505-001 Parameters Module status Measurement: show messurement status Last Service: Service log Entropy, Check sensor: to check the module keyboard. Press the module keys, these 2 items will highlight. Main: If the pulse search is being done. BSR status: show the status for Burst supression classification. There are 4 status: Artefact; Supression; Burst; Invalid (not used) Artefact: artefect whether detected. Diathermy: diathermy whether detected. AD clipped: AD-converted value whether clipped Alg.ver.(Mon), Alg.ver.(Mod), Alg.ver.used: These 3 items show version of entropy algorithm, B40 always use algorithm from monitor. For information on Timeouts etc., see section "NIBP Module." Sensor status Entropy, Check sensor: to check the module keyboard. Press the module keys, these 2 items will highlight. Cable, Sensor: whether detect cable and sensor. Sensor S/N, Sensor Lot: show the sensor information Lead: whether detected each of 3 points of sensor. Imped.: imped. value for each of 3 points of sensor. Imp. Meas: whether Imped. measurement on. Imp.: whether is ok for Imped measurement. 5-35 Document no. 2081505-001 B40/B20 Patient Monitor 6 Set/Test The system contains a watchdog circuitry, which needs refreshment at every 10 seconds. If the refreshment did not occur, the watchdog will reset the main CPU. The purpose of the watchdog is to restart the monitor, if there was a serious malfunction. This feature is useful in two cases: when the main CPU is not able to control the monitor, or when the CPU controls the monitor but detects a serious malfunction. Watchdog tests check proper functionality of the watchdog in various conditions. Watchdog test ensures directly that the watchdog functions properly. Choosing this test prevents the watchdog from refreshing and shows running seconds with an accuracy of 0.1 seconds. The test should have the following result when the watchdog is working properly: The monitor will restart after 10 seconds from the start of the test. In malfunction: ‘>20 s’ is displayed. In this case, the fault is in the watchdog. WD by Overload test ensures the functionality of a feature, where the software controls the monitor, but detects an overload situation in the main CPU. The test should have the following result when the feature is working properly: The monitor will restart after 10 seconds from the start of the test. Factory Reset restores the factory default settings and clears the data memories. Factory reset should be run if the monitor software is replaced or if the Timekeeper battery is replaced. 5-36 Document no. 2081505-001 Service Menu Set / Test Sw version / Unit id ---------------------------- Watchdog WD by Over load Factory Reset Previous Menu ------------------------------------------------------------------------------------------------------------------------------------------------------------------2, 2004-03-15 number:-------------------------------------------- Service Log 7 Service Log Error, event and alarm data is stored in the Service Log. The service log contains information about the occurred monitor errors, events and alarms since the last factory reset or service log reset. Error Log is for selecting the error history view onto the right side of the menu. Error Log shows also some monitoring events like warm and cold starts. Alarm Log is for selecting the alarm history view onto the right side of the menu. Service Log Error Log Alarm Log Scroll Log is for scrolling the error/alarm information on the right side of the menu. Scroll Log Record Log is for recording the service log information onto the recorder. Reset Log Record Log Error History Last errors: 2010-Nov-27 13:50:08 UMBC handler: LOG Msg<Umbc reset> at UmbcChannel. cpp(189) Previous Menu Reset Log is for clearing up the content of the selected service log. This function should be run after a performed maintenance. 5-37 Document no. 2081505-001 B40/B20 Patient Monitor 5-38 Document no. 2081505-001 6 Field replaceable unit Spare part 1 Spare part The following parts will be available as field replaceable spare parts. 1.1 Front cover 1.2 Back cover unit 6-1 Document no. 2081505-001 B40/B20 Patient Monitor 1.3 Frame 1.4 Extension rack 6-2 Document no. 2081505-001 Spare part 1.5 Hemo box 1.6 FRU parts list Item Description Order Code Product 2 FRU B40V2 Tian Ma LED Backlight Display Module 2061869-008 B40 2 FRU B20B30 10.4 INCH COLOR TFT-LCD 2044978-071 B20 2,3 FRU B20B30 10.4 INCH LCD WITH BACKLIGHT DRIVER BOARD 2044978-073 B20 3 FRU B40V2 Tian Ma LED Backlight Display Convert 2061869-009 B40 3 FRU B20B30 LED BACKLIGHT DRIVER BOARD 2044978-072 B20 4 FRU B40B20 EXTERNAL INTERFACE BOARD 2053489-004 B40, B20 6 FRU B40B20 user Interface board 2053489-006 B40, B20 7 FRU B40B20 Speaker 2053489-007 B40, B20 8 FRU B40 Frame cables and mechenical parts 2053489-008 B40 8 FRU B20 FRAME CABLES AND MECHENICAL PARTS 2053518-004 B20 11 FRU B40B20 battery board 2053489-011 B40, B20 12 FRU B40B20 AC/DC unit 2053489-012 B40, B20 13 FRU B40B20 AC inlet 2053489-013 B40, B20 15 FRU B40V2 Handle 2061869-013 B40, B20 21 FRU B40B20 Trim Knob 2053489-021 B40, B20 24 FRU B40B20 Module interface board 2053489-024 B40, B20 43 FRU BX0 B1X5 HEMO NELLCOR (SPO2, TEMP, IBP) (ROHS) FOR NEW DUMP VALVE 2053489-093 B40, B20 44 FRU BX0 B1X5 HEMO MASIMO(SPO2, TEMP, IBP) (ROHS) FOR NEW DUMP VALVE 2053489-092 B40, B20 45 FRU BX0 B1X5 HEMO GE STP (SPO2, TEMP, IBP) FOR NEW DUMP VALVE 2053489-091 B40, B20 46 FRU BX0 B1X5 HEMO GE SPO2 (SPO2 ONLY) FOR NEW DUMP VALVE 2053489-094 B40, B20 47 FRU B40B20 PRINTER (ROSH) 2053489-062 B40, B20 48 FRU B40B20 Recorder connect board (ROSH) 2053489-063 B40, B20 6-3 Document no. 2081505-001 B40/B20 Patient Monitor Item Description Order Code Product 49 FRU B40B20 RAC flex cable 2053489-049 B40, B20 51 FRU B40B20 Adapter (ROSH) 2053489-061 B40, B20 53 FRU B40 ALARM LIGHT 2053489-053 B40 53 FRU B20 ALARM LIGHT 2053518-010 B20 54 FRU B40V2 Battery Cover 2061869-014 B40, B20 64 FRU B40B20B30 BATTERY 2044978-004 B40, B20 69 FRU B40V2 Front Cover without Keypad* 2061869-002 B40 69 FRU B20V2 Front Cover without Keypad* 2061869-056 B20 70 FRU B40V2 Rack Cover 2061869-003 B40, B20 71 FRU B40V2 Power Board 2061869-004 B40, B20 79 FRU B40V3 B20V2 KEYBOARD/MEMBRANE* 2061869-054 B40, B20 - CABLE RS232 0.76 M 2025963-001 B40, B20 NOTE: Please buy this part locally. MPN: CS2N9MF-2.5 NOTE: 2061869-054 FRU B40V3 B20V2 keyboard/Membrane Switch includes multi-languages labels. When service people replace front cover, please order both 2061869-002 and 2061869-054 for B40; both 2061869-056 and 2061869-054 for B20. 1.7 Other parts The following parts have their own instructions accompanying with the FRUs. Description Order Code Product FRU B40V3 B20V2 Software Recovery CD 2061869-052 B40, B20 FRU B40V3B20V2 SOFTWARE RECOVERY CD VSP-C 1.01 SPA01 2061869-085 B40, B20 FRU B40V3B20V2 SOFTWARE RECOVERY DVD VSP-C 1.01 SPA01 2061869-086 AND LATER B40, B20 FRU B40B20 HEMO NIBP FIRMWARE 2053489-077 B40, B20 FRU B20B40 HEMO ECG FIRMWARE 2053489-078 B40, B20 FRU BX0 FILTER POROUS PLASTIC 20U 2061869-088 B40, B20 FRU B20B40 HEMO DUMP VALVE SUB ASSY 2053489-066 B40, B20 FRU B40B20 HEMO LABELS 2053489-080 B40, B20 1.8 DRU Parts More details , please refer to 2062418-001 Service instruction for depot. Description Order No. DRU B40B20 CPU BOARD 2053489-014 DRU B40B20 CPU BATTERY 2053489-005 DRU B40B20B30 Module PCBA ECG 2061869-096 6-4 Document no. 2081505-001 Spare part Description Order No. DRU B40B20B30 Module PCBA STP 2044978-026 DRU B40B20 HEMO COVER 2053489-027 DRU B125B105 HEMO NIBP PUMP 2098800-118 DRU B40B20 GE SPO2 BOARD PCBA 2053489-029 DRU B40B20 HEMO PCBA NIBP 2053489-031 DRU B40B20 HEMO NIBP HOSE CONNECTOR 2053489-032 DRU B40B20 HEMO NIBP TUBE AND CABLES 2053489-033 DRU B40B20 HEMO FLEX CABLES 2053489-035 DRU B40B20 HEMO ECG INPUT UNIT 2061869-095 DRU B40B20 HEMO STP INPUT UNIT 2053489-037 DRU B40B20 MASIMO SPO2 BOARD (ROSH COMPLIANT) 2053489-058 DRU B40B20 NELLCOR SPO2 BOARD (ROSH COMPLIANT) 2053489-057 DRU B40B20 HEMO PCBA TP K10 FOR MASIMO SPO2 2104018-001 DRU B40B20 HEMO PCBA TP FOR NELLCOR SPO2 2053489-042 DRU B40V2/V3 REAR COVER 2061869-005 DRU B20V2 REAR COVER 2061869-055 DRU B40V2/V3 MAIN CHASSIS 2061869-006 DRU B20 MAIN CHASSIS 2053518-011 DRU B40B20 COVER ISOLATION ECG BOARD 2053489-081 6-5 Document no. 2081505-001 B40/B20 Patient Monitor 2 Disassembly WARNING CAUTION A short circuit may cause internal damage. Do not touch any exposed wiring or conductive surface inside, this may cause an electric shock. Perform leakage current measurement whenever service or repair has been done in the monitor. Field repair of the patient monitor is limited to replacing Field Replaceable Units (FRUs). See “1. Spare part” for a detailed list of available FRUs. Attempting a field repair on a printed circuit board or a factory sealed component or assembly could jeopardize the safe and effective operation of the monitor. NOTE: Only a qualified service technician should perform field replacement procedures. NOTE: Perform the checkout procedure described in chapter 3 “4. Maintenance and checkout” after you have disassembled and reassembled the module. 2.1 ESD precautions Refer to 1.2.4. ESD precautionary procedures in Chapter 1. 2.2 Reassembly precautions GE recommends using the new fasteners (screws, washers, etc.) provided in the FRU kits rather than re-using the old fasteners. Some fasteners are not intended to be re-used more than three times. Use only new screws attaching into aluminium. Take advantage of existing thread pattern cut by turning the screw counterclockwise until it drops into the existing thread pattern. • When reassembling the monitor, comply with the given torque [x Nm] for the screws. Note the positions of any wires, cables or connectors. Mark them if necessary to ensure that they are re-assembled correctly. • Save and set aside all hardware for reassembly. 2.3 Required tools • • • Insulated screwdrivers; PH2, PO1, PH1, Hex 5.5 (recommend the length > 65 mm) Insulated Flat blade screwdriver 3 mm Antistatic wristband 2.4 Before disassembly NOTE: Wear a grounded, antistatic wristband when handling PC boards. Electrostatic discharge may damage components on the board. − Turn the monitor off from the On/Off button. − Disconnect the monitor power cord first from the wall outlet and then from the monitor. − Remove the battery. − Disconnect all external cables connected to the monitor. − Detach the E module from the frame. 6-6 Document no. 2081505-001 Disassembly − Detach the monitor from the mount if installed. 2.5 To separate the frame To separate the extension rack, hemo module and multi I/O from the frame 1. Remove the 2 screws from the bottom of the frame NOTE: When place the monitor face down, be careful of the screen and the Trim Knob. 2. Remove the screw beside the monitor. NOTE: When assembly the extension rack, please install this screw first before 2 screw in the bottom 3. From the back side of the monitor, pull the extension rack out of frame, grasp it firmly. 6-7 Document no. 2081505-001 B40/B20 Patient Monitor 4. Open the battery’s cover, use a screwdriver to release the latch by pivoting the screwdriver away from the monitor. NOTE: Using the 3 mm Flat blade screwdriver into the small hole inside, pull the latch to the screen side. 5. Using the screwdriver to pry the module from the monitor at the same time. 6. Pull the hemo module out of the monitor. 7. Use the screwdriver to prize, pulling out the multi I/O at the same time. 6-8 Document no. 2081505-001 Disassembly 2.6 To disassemble the frame - Remove the 2 screws from the back side of the frame, open the cover. - Remove the 4 screws from top of the frame - Remove other 3 screws from the frame - Lift the back cover up To open the back cover of the frame To change the handle After opening the back cover - Remove 2 screws for the handle. 6-9 Document no. 2081505-001 B40/B20 Patient Monitor To remove the front cover To remove the alarm light board - For B40: Remove 4 screws by the side of the frame. - For B20: Remove 4 screws by the side of the frame, and 1 screw on the middile of the frame. - Disconnect the alarm light cable - Disconnect the connection cable - Open and remove the front cover from the frame After opening the front cover - Remove 2 screws, the alarm light board is on the top of the front cover. To remove the user interface board After opening the front cover 6-10 Document no. 2081505-001 Disassembly - Remove 2 screws, the user interface board is on the bottom of the front cover - Disconnect the Trim Knob cable - Disconnect the keyboard/membrane switch cable - Remove 4 screws from beside the display - Disconnect the display inverter cable and remove the screw. - Disconnect the display cable on the bottom side of the screen - Remove the display To separate the display NOTE: When reassemble the display, be careful that no dirt or finger prints are left. To remove the loudspeaker unit After removing the display - Disconnect the cable from CPU board - Remove 2 screws for the loudspeaker To remove the ACDC unit 6-11 Document no. 2081505-001 B40/B20 Patient Monitor - Remove screw - Disconnect the cable to power board. - Disconnect the cable to ethenet board - Slightly lift the front of the ACDC unit, push the whole ACDC unit to top direction. - Lift the unit up - Disconnect the cable to power board - Disconnect all the rest cables from the CPU board - Separate the whole unit from the frame - Remove 3 screws for ethenet board - Lift up the ethenet board from the frame To remove the ethenet board To remove the CPU board 6-12 Document no. 2081505-001 Disassembly - Remove 4 screws for the CPU Board NOTE: Replace the CPU battery every 5 years. - Carefully disconnect the CPU board from the power board, and remove it out. - Remove the rest 7 screws off - Carefully disconnect the CPU board from the power board, and remove it out first - Remove the battery board out of the frame - Remove the 2 screws and separate the board from the unit To remove the power board and battery board To remove the inverter board To remove the ACDC board 6-13 Document no. 2081505-001 B40/B20 Patient Monitor - Remove the four screws in ACDC board - Disconnect the cable from the AC inlet - Lift up ACDC - Disconnect the cable from from the module interface board - Remove 2 screws and then remove the module interface board To remove the module interface board 6-14 Document no. 2081505-001 Disassembly 2.7 To disassemble the extended rack and the recorder To remove the recorder - Use the flat blade screwdriver to press the lock through the hole inside. - Pulling the recorder out at the same time - Remove the 5 screws on the box - Lift the cover up To open the extended rack box To remove the rac flex cable 6-15 Document no. 2081505-001 B40/B20 Patient Monitor - Remove the three screws, then remove off the board - Remove the paper from the recorder - Remove 2 screws inside the recorder To disassemble the recorder Reassemble the module in reverse order. 6-16 Document no. 2081505-001 Disassembly 2.8 Handling and storage of display component Handling of display component and protective window If the Display component surface becomes dusty, wipe it gently with absorbent cotton, chamois or other soft material. If necessary, breathe onto the display surface and wipe immediately. The display surface may also be cleaned using a small amount of normal hexane. Do not use acetone, toluene or alcohol because they cause chemical damage to the polarizer. 1. Wipe off saliva and water drops as soon as possible. Their prolonged contact with the polarizer cause deformations and color fading. 2. Do not open the component case because internal circuits are sensitive to electrostatic discharges. Placing a spare part Display component or a display shield into use: Peel off the protective film slowly (in more than 10 seconds) from the display or protective window surface. Fast peeling may generate enough static electricity to destroy the Display component. Storaging an display component as a spare part for a long period 1. Store the display in a dark place. Do not expose it to sunlight or fluorescent light. Keep the temperature between -30 °C and 85 °C / -22 °F and 185 °F at 5% to 95% humidity. 2. The polarizer surface should not come into contact with any other object. It is recommended that the display unit is stored in the container in which it was originally shipped. 2.9 To replace the fuses Pull out the fuse holder under the mains connector at the back of the monitor. Replace the fuses with fuses of exactly the same type and rating. To change the fuses - Use the screwdraft to press the fuse holder - Prize and pull the holder out - Replace the fuse with correct type and rating 2.10 To download the software Refer to the instruction in Software’s FRU, follow the instruction for downloading service software. 6-17 Document no. 2081505-001 B40/B20 Patient Monitor 6-18 Document no. 2081505-001 7 E-miniC Module Product overview 1 Product overview 1.1 Introduction This chapter provides information for the maintenance and service of the E-miniC module. The module is a single-width plug-in module designed for use with the GE modular monitors. The E-miniC module provides airway measurements. Letter C in the module name stands for CO2. NOTE: E-miniC is intended for patients weighing over 5kg (11lb). 2 3 1 4 Figure 1 Airway gases setup with E-MiniC Module 1. Module for measuring airway gases 2. Anesthesia gas sampling line 3. Airway adapter with sampling line connector 4. Sampling line connector Monitor software compatibility The E-miniC module, is designed for use with VSP-B or later versions respectively. Equipment safety symbols - When displayed on the E-miniC module, indicates that airway gases should be calibrated every six months in normal use and every two months in continuous use. 1.2 Measurement principle 1.2.1 CO2 measurement E-MiniC is a side stream gas analyzer, measuring real time concentrations of CO2. It is a non dispersive infrared analyzer that measures absorption of the gas sample using an optical narrow band filter. 7-1 Document no. 2081505-001 B40/B20 Patient Monitor The infrared radiation detector is thermopile. Concentration of CO2 is calculated from absorption measured at 4.2 to 4.3 μm. Figure 2 Absorbance of CO2 1.3 Main components − Gas sampling system − MiniC measuring unit − CPU board 7-2 Document no. 2081505-001 Product overview 1.3.1 Gas sampling system The sampling system draws a gas sample to the analyzer at a fixed rate. The gas sampling system samples the measured air to the module, and removes water and impurities from it. A sampling line is connected to the water trap. The pump draws gas through the sampling line to the gas measuring unit. After the measurement, the gas is exhausted from the sample gas out connector. The sample flow is nominally 150 ml/min. Figure 3 MiniC block diagram Mini D-fendTM The sample is drawn through the sampling line. The gas then enters the module through the water trap, where it is divided into two flows, a main flow and a side flow. The main flow goes into the analyzer. This flow is separated from the patient side by a hydrophobic filter. The side flow creates a slight subatmospheric pressure within the Mini D-fend water trap which causes fluid removed by the hydrophobic filter to collect in the bottle. Zero valve The main flow passes through a magnetic valve before proceeding to the analyzer. This valve is activated to establish the zero point for the MiniC measuring unit. When the valve is activated, room air is drawn through a filter into the internal system and the gas sensor. 7-3 Document no. 2081505-001 B40/B20 Patient Monitor NafionTM tube 1) A Nafion tube is used between the water trap and the zero valve to balance the sample gas humidity with that of ambient air. The tube prevents errors caused by the effect of water vapor on gas partial pressure when humid gases are measured after calibration with dry gases. Gas analyzers After the zero valve and Nafion tube, the gas passes through the miniC sensor. Sample flow differential pressure transducer The sample flow differential pressure transducer measures pressure drop across a restrictor and calculates the sample flow from the pressure difference. Working pressure transducer The working pressure transducer measures differential pressure between the tubing and ambient air near the miniC sensor. The following messages are based on the obtained pressure values: ‘Sample line blocked’, ‘Check D-fend’, ‘Replace D-fend’, and ‘Check sample gas outlet’. Absolute pressure transducer The absolute pressure transducer measures real-time ambient pressure. Sampling pump The gas sampling pump is a membrane pump run by a DC-motor. Sample flow is measured with a differential pressure transducer across a known restriction. The motor is automatically controlled to maintain a constant flow even when the D-fend water trap ages and starts to get occluded. It also enables the use of sample tubes with varying lengths and diameters. NOTE: In no occasion is the flow reversed towards the patient. Side flow Main flow Mini D-fend Filter Room air Sample line Zero valve MiniC Diff. Pressure Sensor Abs. Pressure Sensor Pump Gas out Figure 4 1. 1) Gas tubing layout Nafion is a trademark of Perma Pure Inc. 7-4 Document no. 2081505-001 minic_fem_gastubing.vsd Diff. Pressure Sensor Product overview 1.3.2 MiniC sensor The miniC sensor is a non-dispersive infrared analyzer measuring absorption of the gas sample at 4.2 to 4.3 μm infrared wavelength, which is selected using an optical narrow band filter. The IR lamp is a filament surrounded by thermal isolation. There is a hole in the isolation, passing the radiation to a conical measuring chamber with 3 mm length. From the sample chamber, the radiation goes into a thermopile detector with an optical filter in front of it. The temperature sensor measures the miniC measuring unit's temperature and it is used for temperature compensation. The miniC sensor is connected to the CPU board and they can’t be replaced separately. Figure 5 MiniC sensor 1.3.3 CPU board The CPU board contains a processor, memories and all the analog signal processing needed. A MiniC measuring unit is attached to the board with a flexible PCB. Also supply voltage and an RS485 serial channel are connected to the CPU board using another flexible PCB. Analog signals (CO2, temperature, absolute and differential pressures and lamp current signals) are fed to the 16-bit A/D converter. The processor controls the A/D converter and calculates the CO2 percentage and respiration rate from this data. The processor controls sample flow by adjusting the pump voltage based on the differential pressure signal. The processor also controls the current of the IR source and keeps it constant. Calibration data is stored on the EEPROM. 7-5 Document no. 2081505-001 B40/B20 Patient Monitor 1.4 Connectors and signals Table 1 Module bus connector pin description Module bus connector 13 25 1 14 Pin No. I/O Signal 1 I RESET RS485 2 I -15 VDC (not used) 3 I +15 VDIRTY 4 I +15VDC (not used) 5 I/O -DATA RS485 6 I/O DATA RS485 7 8 7-6 Document no. 2081505-001 Ground and Shield I -RESET RS485 9 n/c 10 n/c 11 n/c 12 n/c 13 Ground and Shield 14 I +24/+32 VDIRTY depends on power supply (not used) 15 I Ground DIRTY 16 n/c 17 n/c 18 n/c 19 n/c 20 I GASFR (not used) 21 I CTSD (not used) 22 I TXDD (not used) 23 O RXDD (not used) 24 I +5 VDC (not used) 25 I +5 VDC DIRTY (not used) Maintenance and checkout 2 Maintenance and checkout To help ensure the equipment remains in proper operational and functional order, adhere to a good maintenance schedule. Corrective maintenance Service personnel should perform the following checkout procedure after any corrective maintenance, before taking the module back into clinical use: Performed service activity Required checkout procedure Visual inspections Functional check Front cover replaced All steps Not applicable Mini D-fend O-rings All steps Check "Gas sampling system leak test" and "Sample flow check" Module casing opened either for troubleshooting purpose or for replacing any of the internal parts. All steps All steps Planned maintenance Service personnel shall perform the following checkout procedure completely every 12 months after installation: 1. "Replacement of planned maintenance parts" 2. "Visual inspections" 3. "Functional checkout" The instructions include a check form ("APPENDIX D") which should be filled in when performing the procedures. The symbol in the instructions means that the check form should be signed after performing the procedure. 2.1 Replacement of planned maintenance parts 2.1.1 Required parts Replace the following parts that wear in use at the recommended interval. Part Number Description Pieces Replacement interval 733382-HEL Nafion Tube 1 Once a year 656565 Mini D-fendTM O-ring 2 Once a year M1011471 Zero valve air filter 1 Once every 3 years 7-7 Document no. 2081505-001 B40/B20 Patient Monitor It is also recommended to replace the Mini D-fend water trap and the gas sampling line as part of the planned maintenance procedure. NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. 2.1.2 Replacement procedure Replace the specified planned maintenance parts according to the following procedure. Refer to “5. Disassembly and reassembly” for more information. 1. Replace the zero valve air filter once every 3 years. • • Use a small flat blade screwdriver to pull the old zero line air filter. Attach a new zero line air filter into place. Zero line air filter 2. Nafion tube Replace the special tube (Nafion) and check the condition of the internal tubing. • Replace the 300 mm nafion tube in the sample gas line between the Mini D-fend water trap and the zero valve unit. • Check that the tubing inside the module is not contaminated. Any contamination inside the tubing may indicate that the valve or sensor is contaminated too. This can increase a risk of faulty operation in valve or sensor. The gas sensor is not possible to clean in the field. Therefore, replace the whole miniC assembly with a new one. NOTE: The nafion tube do not include the silicon fittings they connect to. Use the original silicon fittings unless they are not damaged or leaking. 3. Replace the Mini D-fend O-rings: • • Detach the Mini D-fend. • Set the new rubber O-rings into place and attach a new Mini D-fend. 7-8 Document no. 2081505-001 Detach the old rubber O-rings that are around the metal Mini D-fend connectors e.g. using a small flat blade screwdriver. Pay special attention not to scratch the metal Mini D-fend connectors and thus causing leaking. Maintenance and checkout 2.2 Visual inspections Detach the module from the module slot and check that: • • • • • the front cover panel is intact all connectors are intact and attached properly. the module box and latch are intact the metal D-fend connectors and the D-fend O-rings are clean and intact the module and the applied parts are clean The cleaning precautions, cleaning requirements, cleaning procedures, and recommended cleaning solutions for the monitor are described in the “B40 User’s Guide”. For details about cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the accessory package. 2.3 Functional checkout 2.3.1 Test setup Required tools − A barometer − A mass flowmeter for measuring air flow, minimum measurement range 100-300 ml/min, accuracy 5% or better in the 100-300ml/min range. − P/N: 755534-HEL Calibration Gas Regulator − P/N M1006864 Calibration Gas Regulator, (US only) − P/N: 755580 Calibration gas 5% CO2 and air, package of 4 cans − P/N 755587 QUICK CAL calibration gas, (US only) − 3 m / 10 ft gas sampling line − A pressure manometer with either an integrated or a separate pressure pump NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. Connections Disconnect the module from the monitor, if connected. Monitor configuration 1. Configure the CO2 waveform field to the monitor screen. 2. check the CO2 unit: Airway Gas > CO2 Setup > Unit, make sure the unit is mmHg. 2.3.2 Procedure 1. Gas sampling system leak test Check the gas sampling system for possible leakages. • • Disconnect the module from the monitor. Connect a new Mini D-fend water trap to the module. 7-9 Document no. 2081505-001 B40/B20 Patient Monitor • • Connect a new gas sampling line to the sampling line connector in the water trap. • • Block the "Sample Gas Out" connector. • Check that the pressure reading does not drop more than 6 mmHg during 1 minute. Connect the other end of the gas sampling line to a pressure manometer and a pressure pump. Pump 100 mmHg ± 20 mmHg pressure to the gas sampling system. Let the pressure stabilize for approximately 10 seconds. NOTE: The gas module shall be disconnected from the monitor during the leak test. 2. Sample flow check Check the sample flow rate. NOTE: Gas measurement is not available during the first 1 minute after the module is connected due to warm-up. Message 'Calibrating gas sensor' is shown in the waveform field. Wait until warm-up is completed before proceeding with the next steps. • • • • Connect the module with the gas sampling line to the monitor. Connect the gas sampling line to the sampling line connector in the water trap. Connect the other end of the gas sampling line to a flowmeter. Check the sample flow rate reading from the flowmeter. The flow rate shall be within the specification limit 150 ± 25 ml/min. NOTE: Readjustment is needed, if the measured value is not within the specification limits. Adjust the sample gas flow rate according to the instructions in section “3.1. Sample flow rate adjustment”. NOTE: If the sampling pump is noisy it indicates possible problems with motor bearing. In this case it is recommended to replace the noisy sampling pump with a new one. 3. Zero valve operation Test the zero valve functionality. • • Connect the gas regulator to the calibration gas container. • Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Parameters > Gas Unit > Gases. • • • • Start feeding calibration gas. Wait until the CO2 value rises to approximately 5%. 7-10 Document no. 2081505-001 Connect the end of the gas sampling line to the regulator on the gas container. Leave the regulator overflow port open to room air. Open the zero valve to room air by selecting Zero valve ctrl. Check that the CO2 values drop back near 0%. Stop feeding the calibration gas. Maintenance and checkout • Turn the zero valve back to the normal measurement position by selecting ZERO to MEAS. 4. Gas calibration Perform gas calibration according to the instructions in section “3.2. Gas calibration”. 5. Ambient pressure Use a barometer to check the operation of the absolute pressure sensor. • Check that the ambient pressure value shown in the Gas service menu does not differ more than ± 13.33 mbar from the value shown by the barometer. 6. Occlusion detection • • Block the tip of the sampling line by your finger. Check that message 'Sample line blocked' appears to the parameter window within 30 seconds. 7. Air leak detection • • Detach the Mini D-fend water trap. Check that message 'Check D-Fend' appears to the parameter window within 30 seconds. 8. Airway gases • • Breathe a minimum of 5 times to the tip of the sampling line. Check that a normal CO2 waveform appears to the waveform field and the EtCO2 and FiCO2 values are updated to the parameter window. 9. Apnea detection • • Stop breathing to the gas sampling line. Check that an 'Apnea' alarm appears to the message field within 30 seconds. 7-11 Document no. 2081505-001 B40/B20 Patient Monitor 2.3.3 Test completion • Fill in all necessary documents. 7-12 Document no. 2081505-001 Calibration and adjustments 3 Calibration and adjustments 3.1 Sample flow rate adjustment Sample flow rate shall be adjusted: • if the sample flow rate check in section “2.3. Functional checkout” failed. 3.1.1 Calibration setup Required tools − A mass flowmeter for measuring air flow, minimum measurement range 100-300 ml/min, accuracy 5% or better in the 100-300 ml/min range. − 3 m / 10 ft Gas sampling line NOTE: See the supplies and accessories document delivered with the manual for compatible accessories NOTE: Use only accurate, properly maintained, calibrated and traceable calibration tools for the parameter calibration to ensure measurement accuracy. NOTE: Refer to the flowmeter documentation for user instructions. Connections • Ensure that the module is connected to the monitor. • Ensure that you have a new Mini D-fend water trap in use. • Connect a new gas sampling line to the sampling line connector in the water trap. • Connect the other end of the gas sampling line to the flow meter. NOTE: Before checking or adjusting the sample flow rate, make sure there is no leakage in the sampling system. 3.1.2 Sample flow rate adjustment 1. Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Parameters > Gas Unit > Gases. 2. Select Sample gain adj 3. Adjust the sample flow to the nominal value 150 ml/min by using the Trim knob to scroll gain value up-down: • To decrease the sample flow rate measured by the flow meter by approximately 7.5 ml/min, add gain value by 0.05. • To increase the sample flow rate measured by the flow meter by approximately 7.5 ml/min, lower the gain value by 0.05. 4. Press Trim Knob to confirm the gain adjustment. Wait until the sample flow value shown returns near to the nominal value 150 ml/min and then check the actual measured flow rate from the flow meter. 5. Repeat steps 3 and 4 until the flow meter shows a 150 ± 25 ml/min flow rate. NOTE: Adjust the flow rate according to the reading in the flow meter. The flow rate reading in the Gases menu is measured by the internal electronics and settles always back to the nominal 150 ml /min independent on the real flow rate. 7-13 Document no. 2081505-001 B40/B20 Patient Monitor 3.2 Gas calibration Gas calibration shall be performed: • • each time planned maintenance is performed. each time corrective maintenance is performed. NOTE: Gas calibration is a normal user action. Refer to the monitor user’s guide for the recommendation for gas calibration interval in clinical use. 3.2.1 Calibration setup Required tools − P/N: 755534-HEL Calibration Gas Regulator − P/N M1006864 Calibration Gas Regulator, (US only) − P/N: 755580 Calibration gas 5% CO2 and air, package of 4 cans − 3 m / 10 ft Gas sampling line − P/N 755587 QUICK CAL calibration gas, (US only) NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. NOTE: Use only the specified GE Healthcare calibration gas for the gas calibration to ensure measurement accuracy. Do not use any other calibration gases. Check the calibration gas container's labelling to ensure that the calibration gas has not expired. NOTE: Ensure that the gas regulator is functioning properly before gas calibration. Refer to the gas regulator's "Instructions for Use" letter for the annual maintenance instructions. Connections 1. Ensure that the module is connected to the monitor. 2. Ensure that you have a new Mini D-fend water trap in use. 3. Connect the gas regulator to the calibration gas container. 4. Connect a new gas sampling line to the sampling line connector in the water trap. 5. Connect the other end of the gas sampling line to the regulator on the gas container. Leave the regulator overflow port open to room air. Figure 6 7-14 Document no. 2081505-001 Connecting a gas regulator to the calibration gas container and connecting a sampling line to the gas regulator. Calibration and adjustments 3.2.2 Procedure NOTE: Gas calibration is not available during the first 5 minutes after the module is connected. For maximum accuracy, let the monitor to warm up for 30 minutes before starting calibration. NOTE: Gas calibration is not available during a 'Sample line blocked', 'Check D-Fend' and 'Check sample gas out’ alarm condition. Resolve the alarm condition before starting calibration. 1. Enter to gas calibration menu: Airway Gas > Gas Calibration 2. The monitor will start automatic zeroing of the gas sensors. Wait until the message 'Zeroing' is replaced by a message 'Zero OK' message. 3. Open the regulator after a message 'Feed gas' is shown. The measured gas concentrations are shown in real-time in the gas calibration menu. Continue feeding the calibration gas until the measured gas concentrations are stabilized and a message 'Adjust' is shown. Close the regulator. 4. Use the Trim Knob to adjust the gas readings shown in the Calibration menu to match with the gas readings in the labelling of the calibration gas container. Press Trim Knob to accept the adjusted values when the gas readings match each other. 5. Wait until a message 'Ok' is shown. NOTE: A message 'Zero error' is shown in case the zeroing fails. NOTE: A message 'Calibr.error' is shown, if you do not start feeding gas within 1 minute after the automatic zeroing is completed, or if the calibration fails due to too large gain adjustment. NOTE: If zeroing or calibration failed, select the Recalibrate to restart the calibration procedure from the beginning. 7-15 Document no. 2081505-001 B40/B20 Patient Monitor 4 Troubleshooting The problems and solutions in this chapter represent only a few of the faults that you may encounter and are not intended to cover every possible problem that may occur. This chapter focuses on troubleshooting technical problems. For clinical and operation issues refer to the "User’s Guide" for troubleshooting. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” each time after you have opened the module casing. If the problem remains, call technical support for service. To ensure accurate problem solving, please be prepared to provide the following information: • • • • Problem description and the troubleshooting done so far. Device information. Error messages displayed, if any. Other information, as requested. 4.1 Visual inspection Before beginning any detailed troubleshooting, complete thorough visual inspection to be sure that: • • • • • • the front cover is intact • • Check that the module in doubt is compatible with the monitor. the water trap connection and disconnection functions properly all connectors are intact, clean and are attached properly the module box and latch are intact the metal D-fend connectors and the D-fend O-rings are clean and intact Check if there are any messages shown in the message field. Find the possible cause and solution from the “Messages” section in "User’s Guide" . Check to the accessories used with the module. If in doubt, replace the accessories with known good ones. If in doubt of having any loose parts or cable connections inside the module, detach the module box by removing the four screws from the back of the module and check that: • • • • • all screws are tightened properly all cables are connected properly tubes are not pinched and there are no sharp bends on them all tubes are connected properly there are no loose objects inside the module 7-16 Document no. 2081505-001 Troubleshooting 4.2 Troubleshooting checklist The following simple troubleshooting hints may help you to localize and isolate a functional problem to the correct unit. Ensure that the monitor is turned on and the module is connected. 4.2.1 Gas sampling system troubleshooting • Faults which can occur in the sampling system are: leaks or blockages in the tubing, failure of the sampling pump or the magnetic valves, or diminishing of the flow rates because of dirt or other matter accumulating in the internal tubing. • Whenever suspecting the sampling system and always after having done any work on the sampling system, check the sampling system for leakages and check the flow rate. • The D-fend Pro water trap should be replaced, when the 'Replace D-Fend' message appears. • If any liquid has entered the IR sensor due to water trap filter failure, contact GE Healthcare service. • If the sampling pump is noisy it indicates possible problems with motor bearing. In this case it is recommended to replace the noisy sampling pump with a new one. • Check that the tubing inside the module is not contaminated. Any contamination inside the tubing may indicate that the valve or sensor is contaminated, too. This can increase a risk of faulty operation in valve or sensor. The gas sensor is not possible to clean in the field. Therefore, replace the whole miniCO2 assembly with a new one. NOTE: All internal tubes are mechanically fragile. Sharp bends may cause leaks and occlusions. 4.2.2 MiniC sensor troubleshooting • The CPU contains factory calibration data for the miniC sensor. Thus, they can't be replaced separately in the field. In case of failure, the complete miniCO2 assembly should be replaced. If the problem remains, contact service. 7-17 Document no. 2081505-001 B40/B20 Patient Monitor 4.3 Troubleshooting charts Problem Cause What to do too low ETCO2 value • sudden decrease in circulation • pulmonary embolism • hyperventilation • very large dead-space • large shunting • leak in sampling system • calibration error • high by-pass flow from ventilator • check all connections • check calibration too high ETCO2 • hypoventilation • increased metabolism • D-fend contaminated • calibration error • change D-fend • check calibration waveform clipped • incorrect scaling • change scale no response to breathing • apnea • (disconnection) • sampling line or water trap loose or • check all connections blocked (air leak) ETCO2 over scale >20% Shown until 32%, specified range 0...20% ETCO2>PaCO2 • sample gas outlet blocked • check that outlet is open • abnormally high ETCO2 (permissive • let the module run without a hypercapnia) • CO2 sensor contaminated • D-fend malfunction sampling line until the CO2 sensor has dried out • change D-fend • unit is mmHg or kPa and ETCO2 is close to • change to “wet gas” by using arterial PCO2 install/service menu • “dry gas” as default Abnormally high (or low) response to CO2, or sudden occlusion message. • Pressure transducer failure. Unable to perform gas calibration. • Gas calibration is not available during the • Wait until warm-up is first 5 minute after the module is connected. • Gas calibration is not available during ‘Sample line blocked’, ‘Check D-fend’, and ‘Check sample gas out’ alarm condition. 7-18 Document no. 2081505-001 • Replace miniCO2 assembly. completed. • Resolve the alarm condition before starting calibration. Disassembly and reassembly 5 Disassembly and reassembly 5.1 Disassembly guidelines Field service of the module is limited to replacing the serviceable parts listed below (see also chapter “6. Service parts”). Attempting a field repair on any other parts could jeopardize the safe and effective operation of the module, and void the warranty. NOTE: Only a qualified service technician should perform field replacement procedures. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” after you have disassembled and reassembled the module. 5.1.1 Serviceable parts • The miniCO2 assembly, including the following components (CPU board, mini CO2 sensor, pump unit, tubing unit and zero valve and pressure transducers) • • • • • • • • tubing unit pump unit zero valve air filter Mini D-fend O-rings Nafion tube Mini D-fend mechanical parts listed in the service parts chapter 5.1.2 Service limitations The following parts are not serviceable: • miniC sensor and CPU board can't be replaced separately. NOTE: The CPU contains factory calibration data for the miniC sensor. In case of failure, the complete miniCO2 assembly should be replaced. Attempts to repair or calibrate the unit elsewhere will adversely affect operation of the unit. The information provided in this document is for reference only. 5.1.3 ESD precautions All external connectors of the module are designed with protection from ESD damage. However, if the module requires service, exposed components and assemblies inside are susceptible to ESD damage. This includes human hands, non-ESD protected work stations or improperly grounded test equipment. The following guidelines may not guarantee a 100% static-free workstation, but can greatly reduce the potential for failure of any electronic assemblies being serviced: • Discharge any static charge you may have built up before handling semiconductors or assemblies containing semiconductors. • A grounded, antistatic wristband or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors. • Use properly grounded test equipment. 7-19 Document no. 2081505-001 B40/B20 Patient Monitor • Use a static-free work surface while handling or working on assemblies containing semiconductors. • Do not remove semiconductors or assemblies containing semiconductors from antistatic containers until absolutely necessary. • • • Do not slide semiconductors or electrical/electronic assemblies across any surface. • • Handle all PCB assemblies by their edges. Do not touch semiconductor leads unless absolutely necessary. Semiconductors and electronic assemblies should be stored only in antistatic bags or boxes. Do not flex or twist a circuit board. 5.1.4 Before disassembly • Note the positions of any sampling tubes, wires or cables. Mark them if necessary to ensure that they are reassembled correctly. • Save and set aside all hardware for reassembly. 5.1.5 Tools needed - pozidrive screwdrivers - flat blade screwdriver - pincers - antistatic wristband 5.2 Disassembly and reassembly procedure Disassembling the airway module (see the exploded view of the module in chapter “6. Service parts”: 1. Remove the two screws from the back of the module. 2. While pressing the release latch, pull the module box slowly backwards and remove it from the main body. To reassemble the module, reverse the order of the disassembly steps. Check that: • • • • • all screws are tightened properly all cables are connected properly tubes are not pinched and there are no sharp ends on them all tubes are connected properly there are no loose objects inside the module 7-20 Document no. 2081505-001 Disassembly and reassembly 5.2.1 Replacing the pump unit 1. Detach the front cover of the module by releasing the snaps that hold the front cover to he front chassis unit by using a small flat blade screwdriver. There are 2 snaps on both sides of the module and 1 snap on the top. 2. Remove the module casing 3. Unplug the two tubes from the back of the front chassis. 4. Detach the front chassis from the metal frame by removing the two screws. 5. Unplug the hose of the pump. 6. Disconnect the pump’s cable from the CPU board. 7. Remove the three screws that connect the pump unit to the board. To reassemble the module, reverse the order of the dissassembly steps. Figure 7 Uncovered E-miniC module 5.2.2 Replacing the miniCO2 assembly 1. Detach the front cover of the module by releasing the snaps that hold the front cover to he front chassis unit by using a small flat blade screwdriver. There are 2 snaps on both sides of the module and 1 snap on the top. 2. Remove the module casing 3. Unplug the two tubes from the back of the front chassis. 4. Detach the front chassis from the metal frame by removing the two screws. 5. Detach the miniCO2 assy from the frame plate by removing the three screws. 6. Disconnect the FM board from the miniCO2 assy. To reassemble the module, reverse the order of the dissassembly steps. 7-21 Document no. 2081505-001 B40/B20 Patient Monitor 6 Service parts 6.1 Ordering parts To order parts, Contact GE Healthcare. Contact information is available at www.gehealthcare.com. Make sure you have all necessary information at hand. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” after you have disassembled and reassembled the module. 6.2 Spare parts for E-miniC 7-22 Document no. 2081505-001 Service parts Item Description Order No. Mini D-fend, pkg of 10 pcs 8002174 - MiniCO2 assembly, including the following components (CPU board, mini CO2 sensor, pump unit, tubing unit and zero valve and pressure transducers) M1013204 1 Module Casing, Single M1021035 2 Front Chassis Unit, E-miniC M1027134 3 M-miniCO2, CONNECTOR FOR D-FEND, MINI CO2 8002173 4 O-RING, O-RING, 2.5x1.6, VITON, SHORE70, BLACK 656565 5 Front Cover, USA, E-miniC M1026941 5 Front Cover, E-miniC M1026943 6 Latch M1021039 7 Torsion Spring M1020935 8 Screw-cross rec. c/s h, stzn, m3x6 * * Part is not available from GE. Source locally. 7-23 Document no. 2081505-001 B40/B20 Patient Monitor Item Description Order No. 9 Metal frame 879184 10 Frame, E-miniC M1024360 11 Nafion tubing 300 mm 733382-HEL 12 MAGN-VALVE, N.O. valve, 3/2, 5VDC, 0.55W, includes seal 585714 13 Module Connection Flex, E-miniC M1027744 14 Cross cylinder head screw M3x6 * 15 Pump Unit for miniC M1013716 15 Air filter, M-MiniC M1011471 17 Tubing Unit for miniC M1013717 18 Cross cylinder head screw M3x10 * 19 Screw-cross rec.c/s h,m3x8,acidproof * * Part is not available from GE. Source locally. 7-24 Document no. 2081505-001 8 E-sCO, E-sCAiO and N-CAiO Modules Product overview 1 Product overview 1.1 Introduction This document provides information for the maintenance and service of the E-sCO, E-sCAiO, N-CAiO modules. These modules are single width plug-in modules. The modules provide airway measurements. Letters in the module name stand for: C = CO2 and N2O, O = patient O2, A = anesthetic agents, and i = agent identification Table 1 Options for E-sCO, E-sCAiO and N-CAiO modules Modules Parameters / measurements CO2 N2O O2 E-sCO X X X E-sCAiO X X X X X X N-CAiO X X X X X X Figure 1 (1) Anesthetic Agent ID agents Additional measurements MAC MACage Balance gas Respiration rate X X X X X Airway gases measurement setup E-sCO, E-sCAiO or N-CAiO module (2) Gas sample, gas sampling line connector on the water trap (3) Gas sampling line (4) Gas sampling line connector on the airway adapter; place the connector upwards (5) Airway adapter with sampling line connector (6) Heat and moisture exchanger with filter (HMEF) (optional) 8-1 Document no. 2081505-001 B40/B20 Patient Monitor 1.2 Measurement principle 1.2.1 CO2, N2O, and agent measurement MiniTPX is a side stream gas analyzer, measuring real time concentrations of CO2, N2O, and anesthetic agents (Halothane, Enflurane, Isoflurane, Desflurane, and Sevoflurane). Figure 2 MiniTPX sensor principle Anesthetic agents or mixtures of two anesthetic agents are automatically identified, and concentrations of the identified agents are measured. MiniTPX also detects mixtures of more than two agents and issues an alarm. MiniTPX is a non-dispersive infrared analyzer, measuring absorption of the gas sample at seven infrared wavelengths, which are selected using optical narrow band filters. The infrared radiation detectors are thermopiles. Concentrations of CO2 and N2O are calculated from absorption measured at 3-5 μm. Figure 3 Absorbance of N2O and CO2 8-2 Document no. 2081505-001 Product overview Identification of anesthetic agents and calculation of their concentrations is performed by measuring absorptions at five wavelengths in the 8-9μm band and solving the concentrations from a set of equations. Figure 4 Infrared absorbance of AAs The measuring accuracy is achieved utilizing numerous software compensations. The compensation parameters are determined individually for each MiniTPX during the factory calibration. 1.2.2 O2 measurement The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge configuration. The signal picked up with a differential pressure transducer unit is generated in a measuring cell with a strong magnetic field that is switched on and off at a main frequency of 164 Hz. The output signal is a DC voltage proportional to the O2 concentration difference between the gas to be measured and the air reference. Figure 5 O2 measurement principle 8-3 Document no. 2081505-001 B40/B20 Patient Monitor 1.3 Main components The respiratory modules consist of: • • • • Gas sampling system MiniTPX measuring unit MiniOM measuring unit CPU board 1.3.1 Controls and connectors 1 2 3 4 Figure 6 E-sCAiO, E-sCO, N-CAiO modules (1) D-fend Pro water trap (2) Gas sample, sampling line connector on the water trap (3) Water trap container (4) Gas exhaust, connector for the gas exhaust line (sampling gas out) Connector Module Description D25 connector all modules Module bus connector 1.3.2 Gas sampling system The gas sampling system draws a 120 ml/min sample from the patient's airway to the module. The sampling system also takes a 30 ml/min flow of room air to the oxygen sensor. When the gas sensors are zeroed, room air is taken through the CO2-absorber to the gas sensors instead of the sampled gas from the patient's breathing. The gas sampling line is connected between the patient circuit and the Gas Sample port on the water trap. The water trap protects the sampling system and gas sensors from liquids and dust. The diagram of the gas sampling system is shown in the figure below: 8-4 Document no. 2081505-001 Product overview Figure 7 Gas sampling system The sampling system has a self diagnostics that detects disturbances in the gas flow, reveals the most common reasons for disturbances, such as occluded sampling line or blocked gas exhaust line, and communicates relevant status messages to the patient monitor. The system is designed so that gas the sampled gas will not flow from the sampling line back to the patient circuit. The parts and connections of the sampling system are streamlined for minimal dead spaces and turbulences in gas flows. All gas inputs of the module have dust filters protecting the sampling system and gas sensors. The water trap acts as a dust filter for the sampled gas and the module should always have the water trap connected. NOTE: It is very important to prevent dust from entering the open gas connections during service operations. D-fend Pro(+) water trap The gas sampling line is connected to the input of the water trap where a special membrane passes gases and vapors but stops liquids. The gas flowing through the membrane continues via the main flow connector of the water trap to the module. The main flow is about 90% of the sample flow. Liquids stopped below the membrane are moved to the water container by a side flow that goes through the water container and the water separation membrane before entering the side flow connector of the water trap. Thus, the side flow also is free of liquids when it gets into the module. In the module, the side flow is connected directly to the pump input and it does not enter the gas sensors. NOTE: The water trap acts as a dust filter for the sampling system and gas sensors. Thus, the module should always have the water trap connected. 8-5 Document no. 2081505-001 B40/B20 Patient Monitor Zero valve and CO2 absorber The zero valve is activated during gas sensor zeroing. Room air is drawn through the CO2-absorber and the zero valve to the gas sensors, and the main flow of sample gas is stopped. The zero gas comes to the sensors through the CO2-absorber that chemically absorbs CO2. The side flow of the water trap flows in the gas sampling line even during zeroing. During normal monitoring, the zero valve is not activated and the sampled gas gets through the zero valve to the gas sensors. Figure 8 Absorber Nafion tubes 1 The Nafion tube between the water trap and the zero valve equalizes the humidity of the sampled gas to ambient level. This will prevent calibration errors caused by the difference in humilities in the sampled breathing gas and the totally dry calibration gas. Another Nafion tube is used between the CO2 absorber and the zero valve to prevent condensation of water generated in the CO2 absorber as by-product of CO2-absorption. Gas sensors After the zero valve, the gas flows trough the MiniTPX sensor that measures the concentrations of all gases but oxygen. The oxygen concentration is measured in the MiniOM sensor that has two inputs. One input draws in a part of the main flow and the other draws in room air as reference gas for the O2 measurement. Sample flow differential pressure transducer The module measures total flow at the input of the gas pump and reference flow at the OM reference line. The sample flow is the difference of these two flows. Working pressure transducer The working pressure transducer measures absolute working pressure between the MiniTPX unit and MiniOM unit. It is used for messages: ‘Sample line blocked’, ‘Check D-fend’, ‘Replace D-fend’ and ‘Check sample gas out’. Pneumatics unit The pneumatics unit contains the zero valve, the occlusion valve and the pneumatics block with tubing connections. 1 Nafion is a registered trademark of Perma Pure Inc. 8-6 Document no. 2081505-001 Product overview The zero valve is activated during the zero level calibrations of gas sensors. The occlusion valve is activated when the sampling line or water trap is occluded. With the activated occlusion valve, the gas pump generates maximal suction trough the “side flow” connector of the water trap, thus maximizing the transfer of liquids from the wet side of the water trap to the container. The pneumatics block contains a network of constrictions to divide the sampled gas in correct proportions to different parts in the module. The first branching takes place in the water trap where incoming flow is divided to the “main flow” and “side flow”. The second branching takes place before the MiniOM sensor. The pneumatics block also contains a pneumatic low pass filter between gas sensors and gas pump. The filter consists of constrictions (resistors) and volumes (capacitors) and it attenuates the pressure pulsation generated in the gas pump so that they do not disturb the operation of the gas sensors. Gas pump unit The gas pump is a membrane pump run by a brushless DC-motor. The pump is adjusted so that the sample gas flow is kept close to its nominal value even when the flow resistances in the sampling line of water trap change. The pump is in a plastic enclosure to minimize the operating noise and mechanical vibration of the pump unit. A pneumatic damping chamber is integrated to enclosure to attenuate the pressure pulsation and noise conducted to the gas exhaust port. Pressure measurements The four pressure sensors on the CPU board are used to measure ambient pressure, working pressure of the MiniTPX and MiniOM sensors and pressure of the reference gas flow to the MiniOM sensor. Sample flow control The gas flow in the sampling line is monitored by measuring the gas flow at the input of the gas pump and the reference flow to the oxygen sensor is estimated by measuring the pressure in the reference gas flow branch. The sample flow is calculated by subtracting the reference flow from the total gas flow. A control loop adjusts the rotation speed of the pump motor so that the gas flow is kept close to 120 ml/min. Gas sampling self-diagnostics The sample flow and the vacuum in the sampling system are used for continuous monitoring of the gas sampling system. The vacuum is calculated in real time as difference of the measured ambient and working pressures. 8-7 Document no. 2081505-001 B40/B20 Patient Monitor The self-diagnostics of the gas sampling system sends the following status data to the patient monitor when specific triggering conditions are met: ‘Check D-Fend’, ‘Sample line blocked’, ‘Replace D-Fend’, ‘Check sample gas out’ and ‘Continuous blockage’. The gas pump is stopped when the 'Sample line blocked' has lasted for more than 1 minute. The module automatically restarts the pump to check whether the abnormal situation has been resolved so that normal gas sampling operation is possible. The gas pump repeats 1 minute full pump, 30 seconds pump off when the ‘Continuous blockage’ message is shown. Figure 9 Gas tubing layout 1.3.3 MiniTPX measuring unit The MiniTPX unit is a non dispersive infrared analyzer, measuring the absorption of the gas sample at seven infrared wavelengths, which are selected using optical narrow band filters. The IR source is a micro-machined heating element with an integrated collimator. From the output of the source, the radiation is passed to a flow optimized measuring chamber. From the sample chamber, radiation goes via a specially designed beam splitter to two detectors units, each with four thermopile detectors and integrated optical filters. The miniTPX measuring unit has two CO2 channels for redundancy purposes. A more detailed description of the measuring principle can be found in section “1.2.1. CO2, N2O, and agent measurement”. Each detector unit also measures the unit's temperature. The module CPU uses it for further processing and temperature compensation of the measured raw signals. The miniTPX unit includes an amplifier board with the following functions: • • On-board 5V regulator and 2.5V reference source. • • PWM controlled power for the IR source. Preamplifiers for the eight thermopile detectors and for the two temperature sensors. A 16 channel buffered multiplexer is used to transfer the signals to the CPU board. An EEPROM memory for storing factory calibration coefficients of the sensor. The input to the amplifier board comprises a 7V DC feed and CPU control signals for the PWM, the multiplexer and the EEPROM. When the module starts up, the calibration coefficients are read to the module CPU and then used for calculating the gas concentrations from the raw data received from the sensor multiplexer. 8-8 Document no. 2081505-001 Product overview Figure 10 MiniTPX measuring unit 1.3.4 MiniOM Oxygen sensor The miniOM sensor measures the concentration of Oxygen in the gas sample. The measurement is based on the magnetic properties of oxygen. The sensor measures the sound pressure generated in the air gap of the magnet at the 164 Hz operating frequency. Two microphones are used for detection and the Oxygen concentration is calculated from the RMS value of the difference of the microphone outputs. The measurement principle is described in more detail in section ”1.2.2. O2 measurement”. The sensor consists of the following functional parts • • • • Pneumatic system Amplifier board MiniOM board Magnet The sensor is shown in the picture below. Figure 11 MiniOM oxygen sensor NOTE: The sensor is assembled in the module using flexible suspension to prevent the mechanical vibrations of the gas pump and cooling fan from disturbing the Oxygen measurement. All gas lines to the sensor must also be carefully assembled so that they do not pick up mechanical vibrations of the module mechanics. Pneumatic System The pneumatic system, together with the gas sampling system of the module creates the gas flows and pressures needed for the oxygen measurement and protection of the microphones from excessive pressure. About 30 ml/min flow of sampled gas comes to the In connector on 8-9 Document no. 2081505-001 B40/B20 Patient Monitor the MiniOM sensor. Room air is drawn to the Ref input of MiniOM also at 30 ml/min rate. About 75% of these flows are conducted to a pressure equalization chamber so that only about a 8 ml/min flow of the two gas streams continue into the air gap of the magnet. All the internal gas flows finally get to a volume enclosed by the sensor board and the sensor body, and then flow out through the Out connection of the sensor. Some of the gas channels and flow restrictors are integrated into the preamplifier electronics board utilizing the multi-layer structure of the LTCC (Low Temperature Co-fired Ceramics) circuit board technology. NOTE: It is very important to prevent dust or liquids from getting into the pneumatic circuit of MiniOM and thus, the gas connections should always be closed with a protecting cap when the sensor is not connected to the module pneumatics. Amplifier Board The amplifier board located in the sensor has two electric microphones for the differential detection of pressure pulses generated in the magnet's air gap. The microphone signals are fed to two identical signal conditioning channels with a band-pass filter and a digitally controlled amplifier. The voltage gains of the amplifiers are set during factory calibration so that the responses of the microphone channels match in spite of differences in microphone's sensitivities. The amplifier board also has an amplifier for the thermistor measuring the temperature of the magnet. MiniOM Board The MiniOM board has five functions • • • • • Drive the magnet coil. Convert the microphone and temperature signals into digital format. Filter digitally the microphone signals and perform the RMS-conversion. Communicate digitally with the module CPU. Store factory calibration data in permanent memory and communicate them to the module CPU. The module CPU provides the coil drive and communication enabling signals and also clock signal for MiniOM board. The FPGA takes care of the coil drive and has also back-up clock in case of CPU clock does not work. The FPGA takes care of the A/Dconversions which are performed with a serial controlled SAR A/D-converter. The digital band pass filtering and RMS conversion of the microphone signals is made with FPGA circuit controlled by the VHDL code stored in the circuit. In order to filter out the disturbances caused by acoustic noise, mechanical vibration and amplifier noise, the band pass filters are designed to have as narrow a pass band as possible without slowing down the filter's response to changes in the amplitude of the 164 Hz signal. The FPGA circuit takes care of the digital communication between the miniOM sensor and the module CPU. The factory calibration coefficients of the sensor are stored in an EEPROM memory on the miniOM board. When the module starts up, the calibration coefficients are read to the module CPU and then used for calculating the O2 concentration from the Oxygen raw data received from the sensor. 1.3.5 CPU board The CPU board contains the processor, memories and an A/D-converter that is common to the whole module. 8-10 Document no. 2081505-001 Product overview The CPU board also contains sensors for pressures, the temperature and humidity as well as drivers for valves, the fan and the pump. The module is connected to the module bus through an RS-485 serial channel. +3.3 V +1.8 V VCC_OM mOM Connector +7 V +5 V mTPX Connector Humidity and Temperature sensor ModBus_Data Analog to Digital coversion DCDC Power Supply 11.5…18 V RS-485 Transceiver Module Bus Connector D25 RS-485 Zero Valve Connector 4 Pressure sensors CPU Atmel AT91RM9200 Valve Control Memories Occlusion Valve Connector SDRAM 32 Mbyte NOR FLASH 8 Mbyte Crystal 16 MHz Crystal 32.678 KHz Figure 12 ETC Valve Connector Pump Control Pump Connector Fan Control Fan Connector User Button Connector Signal processing on CPU board 1.3.6 MiniOM board The miniOmM board contains electronics specific to the MiniOM sensor: FPGA circuit, coil drive, A/D-converter etc. It also contains EEPROM memory that stores calibration data of the oxygen measurement. 1.3.7 Main Component Interactions The figure below describes the functionality of the module and the division of tasks between different components. 8-11 Document no. 2081505-001 B40/B20 Patient Monitor Figure 13 Block diagram 8-12 Document no. 2081505-001 Maintenance and checkout 2 Maintenance and checkout To help ensure the equipment remains in proper operational and functional order, adhere to a good maintenance schedule. WARNING WARNING CAUTION Only perform maintenance procedures specifically described in the manual. Planned maintenance should be carried out annually. Failure to implement the recommended maintenance schedule may cause equipment failure and possible health hazards. Do not apply pressurized air to any outlet or tubing connected to the module. NOTE: The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule, unless an Equipment Maintenance Agreement exists. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device. Corrective maintenance Service personnel shall perform the following checkout procedure after any corrective maintenance, before taking the module back into clinical use: Required checkout procedure Performed service activity Visual inspections Functional check (section 3.2) (section 3.3) Front panel replacement All steps OM Reference gas filter assembly All steps Check “2. Sample Flow Rate Check” Module case opened either for troubleshooting purpose or for replacing any of the internal parts. All steps All steps Planned maintenance Service personnel shall perform the following checkout procedure completely every 12 months after installation: 1. 2.1. Replacement of planned maintenance parts 2. 2.2. Visual inspections 3. 2.3. Functional check The instructions include a check form ("APPENDIX E") which should be filled in when performing the procedures. The symbol in the instructions means that the check form should be signed after performing the procedure. 8-13 Document no. 2081505-001 B40/B20 Patient Monitor 2.1 Replacement of planned maintenance parts 2.1.1 Required parts Replace the following parts that wear in use at the recommended interval. Description Pieces Replacement interval Nafion Tube, 230 mm (mainflow) 1 Once a year Nafion tube, 85 mm (zero line) 1 Once every 4 years OM Reference gas filter assembly including O-ring 1 Once a year PM sticker Once a year CO2 absorber 1 Once every 4 years It is also recommended to replace the D-fend Pro water trap, the gas sampling line as part of the planned maintenance procedure. NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. 2.1.2 Planned Maintenance Kits The required planned maintenance parts are included in a PM kit. Part number M1206554 Description Planned Maintenance Kit for E-sCO, E-sCAiO, N-CAiO modules. The PM kit includes the required Nafion tubes, the OM reference gas filter assembly with an O-ring and a PM sticker. NOTE: The PM kit does not include the CO2 absorber. Order it separately. 2.1.3 Replacement procedures Replace the specified planned maintenance parts according to the chapter "Disassembly and reassembly" on page 8-31. 1. Replace the CO2 absorber every 4 years. 2. Replace the special tubes (Nafion) and check the condition of the internal tubing. • Check that the tubing inside the module is not contaminated. Any contamination inside the tubing may indicate that the valves or sensors are contaminated, too. This can increase a risk of faulty operation in valves or sensors. The valves or gas sensors 8-14 Document no. 2081505-001 Maintenance and checkout are not possible to clean in the field. Therefore, if you noticed any contamination in the module tubing, send the module to GE Healthcare for factory service. NOTE: The nafion tubes do not include the silicon fittings they connect to. Use the original silicon fittings unless they are damaged or leaking. 3. Replace the OM reference gas filter assembly. 4. Check that the fan and ventilation hole are not covered in dust. 2.2 Visual inspections Detach the module from the module slot and check that: • • • • • the front cover is intact all connectors are intact and are attached properly the module box and latch are intact the D-fend Pro and its connectors are clean and intact the module and the applied parts are clean The cleaning precautions, cleaning requirements, cleaning procedures, and recommended cleaning solutions for the monitor are described in the monitor user’s manual. For details about cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the accessory package. 8-15 Document no. 2081505-001 B40/B20 Patient Monitor 2.3 Functional check Turn on the monitor. Wait until the normal monitoring screen appears. 2.3.1 Test setup Required tools − A barometer − A mass flowmeter for measuring air flow, minimum measurement range from 0 to 200 ml/min, accuracy 5% or better in the 0 to 200 ml/min range. − P/N: 755534-HEL Calibration Gas Regulator − P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-sCAiO, N-CAiO modules) − P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with E-sCO module) − P/N: M1006864, Calibration Gas Regulator, US only − P/N: 755571-HEL, Calibration Gas, 5% CO2, 54.5% O2, 36.0% N2O, 2.0% DESFLURANE, BAL N2 (with E-sCAiO, N-CAiO modules) US only − P/N: 755587, Calibration Gas, CO2, O2, Balance, 4 cans/pkg (with E-sCO module) US only − D-fend Pro water trap − 3 m / 10 ft anesthesia gas sampling line − Adult D-Lite sensor − A pressure manometer with either an integrated or a separate pressure pump − Forceps NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. Connections • Disconnect the module from the monitor, if connected. Monitor configuration • Configure the CO2, O2, AA, and Flow waveform fields to the monitor screen with adequate priority. 2.3.2 Procedure Mark each task as complete on the checkout form. 1. Gas Sampling System Leak Test NOTE: The gas module shall be disconnected from the monitor during the leak test. Check the gas sampling system for possible leakages. • • • Disconnect the module from the monitor. Detach the module front cover and casing. Block the OM reference tube with the forceps. Correct positioning of the forceps is indicated by the figure below. NOTE: Be careful when attaching the pincers to the tube and avoid stretching the tube. Short pieces of silicone tubing on the forcep jaws can be used to protect the tube from breaks that may appear when the tube is compressed between the jaws. 8-16 Document no. 2081505-001 Maintenance and checkout • • • Connect a new D-fend Pro water trap to the module. • • Block the sample gas out (gas exhaust) connector. • Check that the pressure reading does not drop more than 2 mmHg during 25 seconds. • Release the forceps, and reassemble the module. Make sure that the tubing fits nicely into the module casing. Connect a new gas sampling line to the sampling line connector in the water trap. Connect the other end of the gas sampling line to a pressure manometer and a pressure pump. Pump 80 mmHg ± 20 mmHg pressure to the gas sampling system. Let the pressure stabilize for approximately 10 - 20 seconds. 2. Sample Flow Rate Check Check the sample flow rate. Connect the module to the monitor. NOTE: Gas measurement is not available during the first 1 to 5 minutes after the module is connected due to warming up. A message 'Calibrating Gas Sensor' is shown in the waveform field. Wait until warm-up is completed before proceeding with the next steps. NOTE: The ambient temperature and air pressure influence the flow rate measured by the flow meter. A flow meter, which has been calibrated at 21.11 °C (70 °F) and 760 mmHg (1033 cmH2O), measures the flow rate correctly under the same conditions, i.e. in room temperature at sea level. A flow rate correction as instructed by the manufacturer of the flow meter needs to be performed when measuring flow rate under other conditions, for example in high altitude. • Connect the gas sampling line (3 m / 10 ft with E-sCO, E-sCAiO and N-CAiO) to the sampling line connector. • • Connect the other end of the gas sampling line to a flowmeter. Check the sample flow rate reading from the flowmeter. The flow rate shall be within the specification limit 120 ± 20 ml/min. NOTE: Readjustment is needed, if the measured value is not within the specification limit. Adjust the sample gas flow rate according to the instructions in section “3.1.2. Sample Flow Rate Adjustment”. 8-17 Document no. 2081505-001 B40/B20 Patient Monitor 3. Reference Gas Flow Rate Check Check the flow rate in reference gas inlet: • • • Connect the module to the monitor. • Check that the Reference Flow is within the following range: 10 - 50 ml/min with E-sCO, E-sCAiO and N-CAiO modules • • Detach the water trap. Leave the other end of the gas sampling line open to room air. Connect the flowmeter to the OM reference gas inlet on the side of the module with a piece of tubing. Attach the front cover. 4. Fan • • Check that the gas module's fan is running behind the D-fend Pro water trap. Attach the water trap 5. Zero Valve Operation Test the zero valve functionality: • • Connect the gas regulator to the calibration gas container. • Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Parameters > Gas Unit > Gases. • Start feeding the specified calibration gas. Wait until the gas values shown in the Gas calibration menu rise approximately to the level indicated in the labelling of the calibration gas container. Connect the end of the gas sampling line to the regulator on the gas container. Leave the regulator overflow port open to room air. NOTE: The gas values in the gas service menu is in percentages (%). • • • • Open the zero valve to room air by selecting Zero valve ctrl. Check that the CO2, N2O and anesthesia agent values drop back near 0% and the O2 reading near 21% (room air). Stop feeding the calibration gas. Turn the zero valve back to the normal measurement position by selecting ZERO to MEAS. (measurement position). 6. Gas Calibration Perform gas calibration according to the instructions in section “3.2. Gas Calibration”. 8-18 Document no. 2081505-001 Maintenance and checkout 7. Agent Identification NOTE: Perform this test only for E-sCAiO and N-CAiO modules. Check agent ID unreliability: • • Feed the specified calibration gas for at least 30 seconds. Check that the anesthesia agent is identified as Desflurane and the ID unrel. value (=agent ID unreliability) shown in the gas service menu is lower than 75. If the value is higher, repeat the gas calibration and check the value again. 8. Ambient Pressure Use a barometer to check the operation of the absolute pressure sensor. • Check that the ambient pressure value shown in the gas service menu does not differ more than ± 13.33 mbar from the value shown by the barometer. 9. Occlusion detection • • Block the tip of the sampling line by your finger. Check that a 'Sample line blocked' and ‘Low gas sample flow’ message appears on the screen within 30 seconds. 10. Air Leak detection • • • Detach the D-fend Pro water trap. Check that the message 'Check D-Fend' appears on the screen within 30 seconds. Attach the water trap. 11. Gas exhaust blockage • • Block the gas exhaust connector with your finger. Check that the message 'Check sample gas out' appears on the screen within 30 seconds. 12. Airway Gases • • Breathe a minimum of 5 times to the tip of the sampling line. Check that a normal CO2 waveform appears to the waveform field and the EtCO2 and FiCO2 values are updated on the screen. 8-19 Document no. 2081505-001 B40/B20 Patient Monitor 13. Apnea detection • • Stop breathing to the gas sampling line. Check that an 'Apnea' alarm appears to the message field within 30 seconds. 2.3.3 Test completion • Select Discharge patient to discard any changes made to the monitor configuration during checkout. • • Disconnect and reconnect the module before starting a new case. Fill in all necessary documents. 8-20 Document no. 2081505-001 Calibration and adjustments 3 Calibration and adjustments 3.1 Sample Flow Rate Adjustment Sample flow rate shall be adjusted: • if the sample flow rate check in section “2.3.2. Procedure” failed. 3.1.1 Calibration setup Required tools − A mass flowmeter for measuring air flow, minimum measurement range 0 to 200 ml/min, accuracy 5% or better in the 0 to 200 ml/min range. − 3 m / 10 ft anesthesia gas sampling line. NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. NOTE: Use only accurate, properly maintained, calibrated and traceable calibration tools for the parameter calibration to ensure measurement accuracy. NOTE: If the flow meter unit is not ml/min, it shall be converted to ml/min according to the instructions of the flow meter manufacturer. NOTE: Gas module sample flow rate is calibrated in the factory to ambient air conditions corresponding the flow at the end of 3 m sampling line. Make sure that your meter is also showing the flow at ambient conditions (= ATP). NOTE: Refer to the flowmeter documentation for user instructions. Connections 1. Ensure that the module is connected to the monitor. 2. Ensure that you have a new D-fend Pro water trap in use. 3. Connect a new gas sampling line to the sampling line connector in the water trap. 4. Connect the other end of the gas sampling line to the flow meter. NOTE: Before checking or adjusting the sample flow, make sure there is no leakage in the sampling system. 3.1.2 Sample Flow Rate Adjustment 1. Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) > Service (26-23-8) > Parameters > Gas Unit > Gases. 2. Select Sample gain adj 3. Adjust the sample flow to the nominal value 120 ml/min by increasing or decreasing Sample Flow Gain: • To decrease the sample flow rate measured by the flow meter by approximately 7.5 ml/min, add gain value by 0.05. • To increase the sample flow rate measured by the flow meter by approximately 7.5 ml/min, lower the gain value by 0.05. 4. Press Trim Knob to confirm the gain adjustment. Wait until the sample flow value shown on screen returns near to the nominal value 120 ml/min and then check the actual measured flow rate from the flow meter. 5. Repeat steps 3 and 4 until the flow meter shows a 120 ± 20 ml /min flow rate. 8-21 Document no. 2081505-001 B40/B20 Patient Monitor NOTE: Adjust the flow rate according to the reading in the flow meter. The flow rate reading in the Gases menu is measured by the internal electronics and settles always back to the nominal 120 ml /min independent on the real flow rate. 3.2 Gas Calibration WARNING WARNING Failure in zeroing or calibrating gases might cause inaccurate readings. Since calibration gas contains anesthetic agents, always ensure sufficient ventilation of the room during calibration. Gas calibration shall be performed: • • each time planned maintenance is performed. each time corrective maintenance is performed. NOTE: Gas calibration is a normal user action. Refer to the monitor user’s manual for the recommendation for gas calibration interval in clinical use. 3.2.1 Calibration setup Required tools − P/N: 755534-HEL Calibration Gas Regulator − P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-sCAiO and N-CAiO modules) − P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with E-sCO module) − P/N: M1006864, Calibration Gas Regulator, US only − P/N: 755571-HEL, Calibration Gas, 5% CO2, 54.5% O2, 36.0% N2O, 2.0% DESFLURANE, BAL N2 (with E-sCAiO and N-CAiO modules) US only − P/N: 755587, Calibration Gas, CO2, O2, Balance, 4 cans/pkg (with E-sCO module) US only − 3 m / 10 ft anesthesia gas sampling line NOTE: Use only the specified GE Healthcare calibration gas for the gas calibration to ensure measurement accuracy. Do not use any other calibration gases. Check the calibration gas container's labelling to ensure that the calibration gas has not expired. NOTE: Ensure that the gas regulator is functioning properly before gas calibration. Refer to the gas regulator's "Instructions for Use" letter for the annual maintenance instructions. Connections 1. Ensure that the module is connected to the monitor. 2. Ensure that you have a new D-fend Pro water trap in use. 3. Connect the gas regulator to the calibration gas container. 4. Connect a new gas sampling line to the sampling line connector in the water trap. 5. Connect the other end of the gas sampling line to the regulator on the gas container. Leave the regulator overflow port open to room air. 8-22 Document no. 2081505-001 Calibration and adjustments Figure 14 Connecting a gas regulator to the calibration gas container and connecting a sampling line to the gas regulator. 3.2.2 Procedure NOTE: Gas calibration is not available during the first 5 minutes after the module is connected. For maximum accuracy, let the monitor to warm up for 30 minutes before starting calibration. NOTE: Gas calibration is not available during a 'Sample line blocked', 'Check D-Fend' and 'Check sample gas out’ alarm condition. Resolve the alarm condition before starting calibration. 1. Enter to gas calibration menu: Airway Gas > Gas Calibration 2. The monitor will start automatic zeroing of the gas sensors. Wait until the message 'Zeroing' is replaced by a message 'Zero Ok' for all measured gases. 3. Open the regulator after a message 'Feed gas' is shown for all measured gases. The measured gas concentrations are shown in real-time in the gas calibration menu. Continue feeding the calibration gas until the measured gas concentrations are stabilized and a message 'Adjust' is shown for all measured gases. Close the regulator. 4. Use the Trim Knob to adjust the gas readings shown in the Calibration menu to match with the gas readings in the labelling of the calibration gas container. Press Trim Knob to accept the adjusted values when the gas readings match each other. 5. Wait until a message 'Ok' is shown for all measured gases. NOTE: A message 'Zero Error' is shown in case the zeroing fails. NOTE: A message 'Calibration Error' is shown, if you do not start feeding gas within 1 minute after the automatic zeroing is completed, or if the calibration fails due to too large gain adjustment. NOTE: If zeroing or calibration failed, select the Recalibrate to restart the calibration procedure from the beginning. 8-23 Document no. 2081505-001 B40/B20 Patient Monitor 4 Troubleshooting The problems and solutions in this chapter represent only a few of the faults that you may encounter and are not intended to cover every possible problem that may occur. This chapter focuses on troubleshooting technical problems. Refer also to the troubleshooting hints on the “User’s Guide” for troubleshooting monitoring problems, performance issues and clinical configuration issues. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” each time after you have opened the module casing. If the problem remains, call technical support for service. To ensure accurate problem solving, please be prepared to provide the following information: • • • • Problem description and the troubleshooting done so far. Device information. Error messages displayed, if any. Other information, as requested. 4.1 Visual inspection Before beginning any detailed troubleshooting, complete thorough visual inspection to be sure that: • • • • • • the front cover is intact • • Check that the module in doubt is compatible with the monitor. the water trap connection and disconnection functions properly all connectors are intact, clean and are attached properly the module box and latch are intact the metal D-fend Pro connectors are clean and intact Check if there are any messages shown in the message field. Find the possible cause and solution from the “Messages” section in "User’s Guide" . Check to the accessories used with the module. If in doubt, replace the accessories with known good ones. If in doubt of having any loose parts or cable connections inside the module, detach the module box by removing the four screws from the back of the module and check that: • • • • • all screws are tightened properly all cables are connected properly tubes are not pinched and there are no sharp bends on them all tubes are connected properly there are no loose objects inside the module 8-24 Document no. 2081505-001 Troubleshooting 4.2 Troubleshooting checklist The following simple troubleshooting hints may help you to localize and isolate a functional problem to the correct unit. Ensure that the monitor is turned on and the module is connected. 4.2.1 Gas sampling system troubleshooting • Faults which can occur in the sampling system are: leaks or blockages in the tubing, failure of the sampling pump or the magnetic valves, or diminishing of the flow rates because of dirt or other matter accumulating in the internal tubing or failure of pressure sensors. • Whenever suspecting the sampling system and always after having done any work on the sampling system, check the sampling system for leakages and check the flow rate. • The D-fend Pro water trap should be replaced, when the 'Replace D-Fend' message appears. • If any liquid has entered the MiniTPX measuring unit due to water trap filter failure, contact GE Healthcare service. • Check that the tubing inside the module is not contaminated. Any contamination inside the tubing may indicate that the valves or sensors are contaminated, too. This can increase a risk of faulty operation in valves or sensors. The valves or gas sensors are not possible to clean in the field. Therefore, if you noticed any contamination in the module tubing, send the module to GE Healthcare for factory service. NOTE: All internal tubes are mechanically fragile. Sharp bends may cause leaks and occlusions. 4.2.2 MiniOM Measuring unit troubleshooting • Due to the complicated and sensitive mechanical construction of the oxygen measuring unit, no repairs should be attempted inside the unit. Instead, if the fault has been found in the measuring unit itself, the entire module should be replaced and the faulty module be sent to GE Healthcare for repair. • In cases of no response to O2 or strong drift, check the tubing for loose connections, blockages, and leaks. • • Check also the OM reference gas filter assembly, and replace if needed. If the O2 signal is noisy, check the measurement unit suspension and if the MiniOM tubing has tension. NOTE: Never apply overpressure to the O2 measuring unit, as the pressure transducer may be permanently damaged. 4.2.3 MiniTPX Measuring unit troubleshooting • The MiniTPX measuring unit can only be repaired at the factory. In case of failure, the entire module should be replaced and the faulty module be sent to GE Healthcare for repair. 4.2.4 CPU board troubleshooting • Due to the complexity of the large scale integrated circuitry, there are few faults in the CPU digital electronics that can be located without special equipment. 8-25 Document no. 2081505-001 B40/B20 Patient Monitor • • Check that all connectors and screws are properly installed. In case of failure, the entire module should be replaced and the faulty module be sent to GE Healthcare for repair. 4.3 Service Interface The monitor has a service menu, which is a useful tool to examine monitor functions and troubleshoot it in case a fault occurs. To enter to the service menu see Chapter 5. 8-26 Document no. 2081505-001 Troubleshooting 4.4 Messages 4.4.1 Gas measurement Message Possible causes Water trap is not connected. Check D-Fend Check D-Fend and sample gas out. Wait for Air leak inside the internal tubing. 30 sec and press Normal screen to continue. Sample line blocked Continuous blockage. Check sampling line and D-Fend. Connect the water trap and sampling line to the module. Check the internal tubing for leakages. Replace, if needed. Gas sampling line is blocked. Check the external gas sampling line for blockages. Replace, if needed. Water trap container is full. Empty the water trap container. Water trap is occluded. Replace the water trap. Internal tubing is blocked. Check the internal tubing for blockages. Replace, if needed. Check sampling gas out The sample gas outflow is blocked. Check D-Fend and sample gas out. Wait for 30 sec and press Normal screen to continue. Internal tubing is blocked. Pump failure. Replace D-Fend Possible solutions Check the sample gas out connector in the front panel and the exhaust line for gas return or scavenging for blockages. - If the sample gas is returned to the patient circuit, check that there is no occlusion in the tubing. - If the sample gas outlet is connected to a scavenging system, make sure an open system is used where gas is removed in room pressure. Check the internal tubing for blockages. Check sample pump operation by measuring the sample gas flow rate. Replace pump, if needed. Defective or contaminated D-fend Pro. Replace the D-fend Pro water trap. Occlusion in internal tubing. Check sample and reference flows. Perform a visual check for the internal tubing. Remove the cause for occlusion. Calibrating Gas calibration is in progress. Wait until the calibration is completed successfully. Failure in Agent ID Agent ID has failed. Perform gas calibration. Check agent ID unreliability (see functional check). If it does not help, send the module to GE Healthcare for factory repair. Zeroing Zeroing is in progress. Wait until zeroing is completed successfully. 8-27 Document no. 2081505-001 B40/B20 Patient Monitor Message Possible causes Possible solutions Zero error Autozeroing during the measurement or Check the zero valve operation. Replace in the beginning of the gas calibration the zero absorber and Nafion tube in failed. zero line. If it does not help, send the module to GE Healthcare for factory repair. Calibrating error Feeding the calibration gas was not started within 1 minute after the automatic zeroing was completed. Recalibrate. Calibration was failed due to too large gain adjustment. Recalibrate. Wrong calibration gas is used. Use the specified calibration gas. Over range Measured FiO2 >103% Perform gas calibration. Apnea deactivated Apnea alarm start-up conditions are not Apnea alarm detection is activated after reached. the 3 breaths are detected. Sensor INOP IR Lamp failure. Check miniTPX flex cable connection. Ambient pressure is too high or low. Check the ambient pressure from the Gas Calibrations menu. CPU failure. Replace CPU. No response from the gas module, high Return the module to GE Healthcare for service. temperature inside the module, or EEPROM checksum failure. Incompatible gas module Incompatible gas module detected by the monitor. Check the compatibility of the gas module. Gas measurements removed The module is disconnected. Reconnect the module. Identical gas modules The monitor detects gas measurement from two or more modules. Remove excess modules providing gas measurement. 8-28 Document no. 2081505-001 Troubleshooting Message Possible causes Possible solutions Sensor INOP MiniOM unit failure Check flex cable connection. - temperature - internal supply voltages - other internal failure MiniTPX unit failure - temperature - internal supply voltages - IR lamp failure CPU failure - internal supply voltages - pressure sensor failure - a/d-converter system failure Other failure - Fan failure - Pump failure - Valve (Zero, Occlusion) failure - Check that fan can rotate freely. Check fan, pump or valve wire CO2 reference signal differs too much connection. from CO2 signal Zeroing fails too many times Calibrating gas sensor O2, CO2 and N2O measurements are not Wait until the warm-up is completed. available during the first minute after the module is connected due to warm-up. Anaesthesia agent measurement is not available during the first 5 minutes after the module is connected due to warm-up. Over Scale Incorrect waveform scale for the parameter. The waveform clipped because measured gas concentration exceeds the upper limit of the current scale. Change to the appropriate waveform scale. For detailed instructions refer to the user’s manual. Low gas sample flow Sample flow deviates to less than 80% of the module specific nominal flow value. Check sample flow rate. Adjust, if needed. Gas sampling line, gas output, water trap, or internal tubing is blocked. Check or replace the gas sampling line, water trap, or internal tubing. Pump failure. Replace the pump unit. 8-29 Document no. 2081505-001 B40/B20 Patient Monitor 4.5 Troubleshooting charts 4.5.1 Gas measurement Problem Possible clinical cause Possible technical cause too low EtCO2 value - sudden decrease in circulation - leak in sampling system - check all connections - calibration error - check calibration - pulmonary embolism - - hyperventilation high by-pass flow from ventilator - very large dead-space - large shunting - hypoventilation - D-fend Pro contaminated - change D-fend Pro - increased metabolism - calibration error - check calibration - incorrect scaling - change scale - sampling line or water trap loose or blocked (air leak) - check all connections - sample gas out blocked - check that outlet is open abnormally high EtCO2 - CO2 sensor contaminated - (permissive hypercapnia) - D-fend Pro malfunction - change D-fend Pro - - Dry gas as default - change to Wet gas For detailed instructions refer to the user’s manual. too high EtCO2 waveform clipped no response to breathing EtCO2 overscale >15% (>20%) Shown until 32%, - apnea - disconnection - What to do call service technician specified range 0...15% EtCO2>PaCO2 unit is mmHg or kPa and EtCO2 is close to arterial PCO2 8-30 Document no. 2081505-001 Disassembly and reassembly 5 Disassembly and reassembly 5.1 Disassembly guidelines WARNING WARNING WARNING Disconnect the module from any monitoring system before performing any repair. Always perform Gas Sampling System Leak Test after the module cover is reassembled. Always perform gas calibration after any planned or corrective maintenance. Field service of the module is limited to replacing the serviceable parts listed below (see also chapter “6. Service parts”). Attempting a field repair on any other parts could jeopardize the safe and effective operation of the module, and void the warranty. NOTE: Only a qualified service technician should perform field replacement procedures. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” after you have disassembled and reassembled the module. 5.1.1 Serviceable parts • • • • • • • • CO2 Absorber D-fend Pro Nafion tubes Front chassis unit Pump OM reference filter Latch and spring Mechanical parts listed in chapter “6. Service parts” 5.1.2 Service limitations The following parts are not serviceable: • • MiniOM Measuring unit MiniTPX measuring unit NOTE: Due to the complicated and sensitive mechanical construction of the oxygen measuring unit, no repairs should be attempted inside the unit. Instead, if the fault has been found in the measuring unit itself, the entire module should be replaced and the faulty module be sent to GE Healthcare for repair. NOTE: The MiniTPX measuring unit can only be repaired and calibrated at the factory. In case of failure, the entire module should be replaced and the faulty module be sent to GE Healthcare for repair. 5.1.3 ESD precautions WARNING Protect module from electrostatic discharge. All external connectors of the module are designed with protection from ESD damage. However, if the module requires service, exposed components and assemblies inside are susceptible to ESD damage. This includes human hands, non-ESD protected work stations or 8-31 Document no. 2081505-001 B40/B20 Patient Monitor improperly grounded test equipment. The following guidelines may not guarantee a 100% static-free workstation, but can greatly reduce the potential for failure of any electronic assemblies being serviced: • Discharge any static charge you may have built up before handling semiconductors or assemblies containing semiconductors. • A grounded, antistatic wristband or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors. • • Use properly grounded test equipment. • Do not remove semiconductors or assemblies containing semiconductors from antistatic containers until absolutely necessary. • • • Do not slide semiconductors or electrical/electronic assemblies across any surface. • • Handle all PCB assemblies by their edges. Use a static-free work surface while handling or working on assemblies containing semiconductors. Do not touch semiconductor leads unless absolutely necessary. Semiconductors and electronic assemblies should be stored only in antistatic bags or boxes. Do not flex or twist a circuit board. 5.1.4 Protection from dust WARNING Module must be handled to prevent dust from entering the gas sampling system. The gas sampling system must be protected from dust entering the tubes, valves and other components. In order to achieve this goal, the following measures must be taken: • • • • Have the D-fend Pro water trap always connected to the module. • Remove the protective caps on the gas pump only immediately before assembling it to the module. • Take the CO2-absorber out from the plastic bag only immediately before assembling it to the module. • The clothing of the service person must be such that the dust risk is taken into account. Clean and dust free working environment during all service procedures. Minimize the times with any open connections in the gas sampling system. Always close the open tube connections of the sampling system when not working on the module. 5.1.5 Before disassembly • Note the positions of any sampling tubes, wires or cables. Mark them if necessary to ensure that they are reassembled correctly. • Save and set aside all hardware for reassembly. 8-32 Document no. 2081505-001 Disassembly and reassembly 5.1.6 Required tools - Torx T8 and T10 screwdrivers - flat blade screwdriver - forceps - antistatic wristband 5.2 Disassembly and reassembly procedure Disassembling the module (see the exploded view of the module in chapter “Service parts”: Reassembling the module: reverse the order of the disassembly steps. Check that: • • • • all screws are tightened properly all cables are connected properly tubes are not pinched and there are no sharp bends on them all tubes are connected properly NOTE: Make sure that the Nafion tubes are routed in such a way that they don’t come near the fan, and there is no risk of the fan being obstructed by the tubes. An obstructed fan will result in degraded ventilation inside the module, and a ‘Sensor inop’ message being displayed. • there are no loose objects inside the module 5.2.1 Disassembly workflow Use this workflow diagram to find the simplest way to disassemble the required parts of the module. Follow the arrows from the top down to the required part and disassemble the module by following the steps in between. E-sCO,E-sCAiO or N-CAiO module D-Fend Front cover Module casing Nafion CO2 absorber OM holder Front chassis unit Main flow connector Latch Pump 8-33 Document no. 2081505-001 B40/B20 Patient Monitor 5.2.2 Detaching the Front Cover 1. Remove the D-fend Pro. 2. Release the two snaps on both sides of the module by using a flat blade screwdriver. 3. Detach the front cover. 5.2.3 Detaching the Module Casing 1. Remove the two T10 screws mounting the D25 connector shield. 2. Detach the connector shield. 3. Remove the two Pozidrive screws. 4. Push the latch and pull the module casing. NOTE: When reassembling ensure that the module casing does not damage the conductive sealings on the front chassis unit. 5.2.4 Replacement of Planned Maintenance Parts 1. Carefully remove the main flow nafion tube and every 4th year the shorter zero line nafion tube. NOTE: Remember the route of the tubes and reassemble correctly. NOTE: Make sure that the Nafion tubes are routed in such a way that they don’t come near the fan, and there is no risk of the fan being obstructed by the tubes. An obstructed fan will result in degraded ventilation inside the module, and sensor inop message being displayed. NOTE: The nafion tubes do not include the silicon fittings they are connected to. Use the original silicon fittings unless they are not damaged or leaking. 8-34 Document no. 2081505-001 Disassembly and reassembly 2. Pull out the OM reference filter assembly with forceps. 3. Push the new filter assembly until it is on the same level with the front chassis. 5.2.5 Replacement of CO2 Absorber 1. Lift the CO2 absorber from the slot. 2. Detach the tubes from the absorber. 3. Connect the tubes to a new CO2 absorber and place it to the slot. 5.2.6 Detaching the Latch 1. Pull the latch from the front chassis. NOTE: Remember to detach the front cover first. 2. Remove the spring by squeezing it. 8-35 Document no. 2081505-001 B40/B20 Patient Monitor 5.2.7 Detaching the Front Chassis Unit 1. Remove the two T10 screws. 2. Carefully detach the three tubes. 3. Disconnect the fan cable. 4. Detach the front chassis unit. 5.2.8 Detaching the Main Flow Connector Original Main Flow Connector is required to maintain proper gas flow restriction in the module. When the Front Chassis Unit is replaced move the original connector to the new unit. 8-36 Document no. 2081505-001 Disassembly and reassembly 1. Carefully detach the lock pin holding the main flow connector. 2. Carefully attach the main flow connector to the new front chassis unit. 5.2.9 Detaching the Pump 1. Carefully detach the tube from the pump. 2. Disconnect the pump cable from the CPU board. 3. Lift the pump from the metal brackets. 8-37 Document no. 2081505-001 B40/B20 Patient Monitor 5.2.10 Detaching the OM holder 1. Carefully detach the three tubes from the OM unit. 2. Carefully disconnect the OM flex cable from the CPU board. 3. Carefully pass the flex cable through metal frame. 4. Lift the OM unit with the holder from the metal brackets. 5. Detach the holder. 8-38 Document no. 2081505-001 Service parts 6 Service parts 6.1 Ordering parts To order parts, Contact GE Healthcare. Contact information is available at www.gehealthcare.com. Make sure you have all necessary information at hand. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” after you have disassembled and reassembled the module. 6.1.1 Planned Maintenance Kits The required planned maintenance parts are included in a PM kit. Part number Description M1206554 Planned Maintenance Kit for E-sCO, E-sCAiO, N-CAiO modules. The PM kit includes the required Nafion tubes, OM reference filter assembly with O-ring and PM Sticker. NOTE: The PM kit does not include the CO2 absorber. Order it separately. 6.2 Spare parts for E-sCAiO, E-sCO, N-CAiO 8-39 Document no. 2081505-001 B40/B20 Patient Monitor Item Description Order No. 1 FRU, Front Chassis Unit, E-sCAiO, E-sCO, N-CAiO M1206530 3 Front Chassis Assembly FRU, HW Kit, E-sCO, E-sCAiO, N-CAiO - All Mounting Screws - All Conductive Sealings - Pump Connector Silicone Tube - D25 Connector Shield - Latch - Torsion Spring M1206533 Item Description Order No. 4 FRU, Pump Unit, E-sCO, E-sCAiO, N-CAiO M1206290 5 FRU, OM Holder, E-sCO, E-sCAiO, N-CAiO M1206531 7 CO2 Absorber, E-sCO, E-sCAiO, N-CAiO M1206555 8-40 Document no. 2081505-001 Service parts 6.2.1 Front covers Item Description Order No. 8 FRU, Front Cover, E-sCAiO M1206538 8 FRU, Front Cover, US, E-sCAiO M1206558 8 FRU, Front Cover, E-sCO M1206539 8 FRU, Front Cover, US, E-sCO M1206559 8 FRU, Front Cover, N-CAiO 2074385-001 8 FRU, Front Cover, US, N-CAiO 2074386-001 8-41 Document no. 2081505-001 B40/B20 Patient Monitor 8-42 Document no. 2081505-001 9 E-Entropy Module Product overview 1 Product overview 1.1 Introduction This document provides information for the maintenance and service of the Entropy Module, E-ENTROPY. The Entropy module uses an electroencephalography (EEG) signal, together with spontaneous facial muscular activity with a frontal electromyography (FEMG) signal to measure: • • • Response Entropy (RE) State Entropy (SE) Burst Suppression Ratio (BSR) The Entropy module is responsible for EEG and FEMG signal acquisition, amplification, filtering and digitization, and electrode impedance measurement. All the calculated parameters can be selected on the display and trended. Figure 1 Measurement setup (1) Entropy module (2) Entropy sensor cable (3) GE Entropy sensor (4) Entropy sensor 1.2 Measurement principle Entropy is a measure of irregularity in any signal. During general anesthesia, EEG changes from irregular to more regular patterns when anesthesia deepens. Similarly, FEMG quiets down as the deeper parts of the brain are increasingly saturated with anesthetics. The Entropy Module measures these changes by quantifying the irregularity of EEG and FEMG signals. Entropy parameters and BSR are calculated from EEG and FEMG signals acquired with a sensor which is attached to the patient's forehead. The sensor consists of three electrodes. This referential measurement yields one channel of raw EntrEEG. 9-1 Document no. 2081505-001 B40/B20 Patient Monitor Figure 2 A general view of the cable connections 1.2.1 EntrEEG EEG is a differential voltage signal measured from electrodes attached to the patient’s skin. EEG measures the spontaneous electrical activity of the brain. This electrical activity reflects the state of the brain. In referential measurement, the referential electrode delivers its potential to every channel’s minus-input. The signal is the potential difference between this common reference electrode (electrode #3) and the electrode connected to the plus input (electrode #1). The purpose of the ground electrode (electrode #2) is to reduce common mode noise. The EntrEEG signal is amplified, antialias filtered, digitized and software filtered. After that, the EntrEEG signal is shown on the screen and the RE, SE and BSR characteristics are calculated from it. 1.2.2 FEMG FEMG is an electrical signal originating from facial muscles. The FEMG signal has much broader spectrum than EEG and it overlaps with EEG at low frequencies. The FEMG signal gives its contribution to the RE values (see “RE and SE”). Mains power frequency and its harmonics are digitally filtered away to reject interference noise from power lines. 1.2.3 RE and SE Entropy numbers range from 100 to zero (RE 0-100, SE 0 – 91), correlating to the patient's anesthetic state. High values of Entropy indicate high irregularity of the signal, signifying that the patient is awake. There are two Entropy parameters: the fast-reacting Response Entropy and the more steady and robust State Entropy. State Entropy consists of the entropy of EEG signal calculated up to 32 Hz. Response Entropy includes additional high frequencies up to 47 Hz and, consequently, the fast frontalis EMG (FEMG) signals enable a fast response time. Table 1 Frequency ranges for Entropy calculation Response Entropy, RE 0 < f < 47 Hz State Entropy, SE 0 < f < 32 Hz 9-2 Document no. 2081505-001 Product overview 1.2.4 Impedance measurement The impedance measurement is performed for all leads at the same time and the EntrEEG is stopped for no longer than 5 seconds during the impedance measurement. Differences in the electrode impedance of the electrodes cause common mode noise coupling to the measured signal. To minimize this, the electrode impedance is measured and a warning of an unsatisfactory impedance level is generated when necessary. The impedance of an electrode is measured by applying a known current through the electrode and measuring the voltage drop over the electrode. This way the impedance of a single electrode can be resolved. 1.3 Main components 1.3.1 Controls and connectors Figure 3 Front panel of Entropy Module, E-ENTROPY, and the back of the module Module key Module Description Entropy E-Entropy Opens the Entropy menu on the screen Check Sensor E-Entropy Starts the manual sensor check Connector Module Description Entropy E-Entropy Entropy connector D25 connector E-Entropy Module bus connector 9-3 Document no. 2081505-001 B40/B20 Patient Monitor 1.3.2 Entropy board The Entropy board consists of the following functional sections: • • • Microprocessor for measurement control, and for processing the measurement signal Digital I/O circuit for smart chip communication (the chip is located in the entropy sensor) Serial communication driver for module bus communication Figure 4 9-4 Document no. 2081505-001 Entropy measurement system block diagram IMC Connector Board Amplification & Filtering A/D Converter Patient Isolation Module Bus The serial bus speed to the monitor is 500 kbps and the bus itself is half duplex, i.e. data can be transferred to both directions but only one way at a time. Maintenance and checkout 2 Maintenance and checkout To help ensure the equipment remains in proper operational and functional order, adhere to a good maintenance schedule. Corrective maintenance Service personnel shall perform the following checkout procedure after any corrective maintenance, before taking the module back into clinical use: Required checkout procedure Performed service activity Visual inspections Electrical safety test (section 3.1) (section 3.2) Functional check (section 3.3) Module casing opened either for troubleshooting purpose or for replacing any of the internal parts. All steps All steps All steps Front cover, or an other external part, replaced. All steps Not applicable Not applicable Planned maintenance Service personnel shall perform the following checkout procedure completely every 24 months after installation: 1. Visual inspections 2. Electrical safety tests 3. Functional check 2.1 Visual inspections Detach the module from the module slot and check that: • • • • the front cover is intact all connectors are intact and attached properly the module box and latch are intact the module and the applied parts are clean The cleaning precautions, cleaning requirements, cleaning procedures, and recommended cleaning solutions for the monitor are described in the "User’s Guide". For details about cleaning, disinfecting and sterilizing of the accessories, see the instructions for use in the accessory package. 9-5 Document no. 2081505-001 B40/B20 Patient Monitor 2.2 Electrical safety tests Complete the Electrical safety tests found in the “Electrical Safety Tests”, chapter 3. Perform Patient leakage current tests, Patient (source) leakage current test and Patient (sink) leakage current test. Record the values of the tests on the "Service check form, E-Entropy" on page F-1. 2.3 Functional check Turn the monitor on. Wait until the normal monitoring screen appears. 2.3.1 Test setup Required tools − P/N: N-ES, Simulator for E-ENTROPY − Entropy sensor cable NOTE: See the supplies and accessories document delivered with the manual for compatible accessories. Connections • Ensure that the module is connected to the monitor. Monitor configuration 1. Configure the Entropy waveform field to the monitor screen with adequate priority. 2. Press the Entropy module key in the module to open the Entropy Setup menu and configure: Entr.EEG Scale: 250uV Display format: RE+SE Automatic sensor check: ON 2.3.2 Entropy tests 1. Module and sensor recognition • • Connect the Entropy sensor cable to the module. • Check that a 'No sensor' message is shown in the Entropy parameter window. Check that the Entropy EEG waveform field and related information appears to the screen. 2. Sensor checking • • 9-6 Document no. 2081505-001 Connect the Entropy simulator to the Entropy sensor cable. Check that a 'Checking sensor' message appears after a while and wait until all 3 electrodes are checked. The sensor check is passed, if a green circle with a check mark changes to orange with result text (PASS/FAIL) for each tested electrode is shown. Maintenance and checkout 3. Functional check with a simulator • Check that the EntrEEG waveform and RE & SE values appear on the monitor screen approximately 30 seconds after the sensor check is completed. 2.3.3 Test completion • Fill in all necessary documents, refer to "Service check form, E-Entropy" on page F-1 9-7 Document no. 2081505-001 B40/B20 Patient Monitor 3 Troubleshooting The problems and solutions in this chapter represent only a few of the faults that you may encounter and are not intended to cover every possible problem that may occur. This chapter focuses on troubleshooting technical problems. For clinical and operation issues refer to the "User’s Guide" for troubleshooting. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” each time after you have opened the module casing. If the problem remains, call technical support for service. To ensure accurate problem solving, please be prepared to provide the following information: • • • • Problem description and the troubleshooting done so far. Device information. Error messages displayed, if any. Other information, as requested. 3.1 Visual inspection Before beginning any detailed troubleshooting, complete thorough visual inspection to be sure that: • • • • the front cover is intact • • Check that the module in doubt is compatible with the monitor. all connectors are intact, clean, and attached properly the module box and latch are intact Check if there are any messages shown in the message field. Find the possible cause and solution from the “Messages” section in "User’s Guide" . Check to the accessories used with the module. If in doubt, replace the accessories with known good ones. If in doubt of having any loose parts or cable connections inside the module, detach the module box by removing the two screws from the back of the module and check that: • • • • • screws are tightened properly cables are connected properly the EMC cover is attached properly in the module there are no loose objects inside the module Press a module key. Check that the related menu is opened or activity is started. Inoperatibility may refer to a loose keypad cable or other problem in the module. 9-8 Document no. 2081505-001 Troubleshooting 3.2 Troubleshooting chart Problem Possible causes No EntEEG waveform or Entropy Entropy module is not connected. parameter window on the screen. Possible solutions Connect the Entropy module to the monitor and the Entropy sensor cable to the module. Configure EntEEG waveform or EntEEG waveform or Entropy Entropy parameter to the screen, parameter window is not configured to or increase the priority of the the screen, or the parameter priority is parameter information. set too low. 9-9 Document no. 2081505-001 B40/B20 Patient Monitor 4 Disassembly and reassembly 4.1 Disassembly guidelines Field service of the module is limited to replacing the serviceable parts listed below (see also chapter “5. Service parts”). Attempting a field repair on any other parts could jeopardize the safe and effective operation of the module, and void the warranty. NOTE: Only a qualified service technician should perform field replacement procedures. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” after you have disassembled and reassembled the module. 4.1.1 ESD precautions All external connectors of the module are designed with protection from ESD damage. However, if the module requires service, exposed components and assemblies inside are susceptible to ESD damage. This includes human hands, non-ESD protected work stations or improperly grounded test equipment. The following guidelines may not guarantee a 100% static-free workstation, but can greatly reduce the potential for failure of any electronic assemblies being serviced: • Discharge any static charge you may have built up before handling semiconductors or assemblies containing semiconductors. • A grounded, antistatic wristband or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors. • • Use properly grounded test equipment. • Do not remove semiconductors or assemblies containing semiconductors from antistatic containers until absolutely necessary. • • • Do not slide semiconductors or electrical/electronic assemblies across any surface. • • Handle all PCB assemblies by their edges. Use a static-free work surface while handling or working on assemblies containing semiconductors. Do not touch semiconductor leads unless absolutely necessary. Semiconductors and electronic assemblies should be stored only in antistatic bags or boxes. Do not flex or twist a circuit board. 4.1.2 Before disassembly • • Note the positions of any wires or cables. Mark them if necessary to ensure that they are re-assembled correctly. Save and set aside all hardware for reassembly. 9-10 Document no. 2081505-001 Disassembly and reassembly 4.1.3 Required tools - antistatic wristband - pincers - torx screwdriver, T10 - flat blade screwdriver 4.2 Disassembling and reassembling procedure Disassembling the E-ENTROPY module (see the exploded view of the module in chapter “5. Service parts”): 4.2.1 To replace the front cover 1. Detach the front cover of the module by releasing the snaps that hold the front cover to the front chassis unit by using a small flat blade screwdriver. There are 2 snaps on both sides of the module and 1 snap on the top. 4.2.2 To replace the Entropy board 1. Remove the two screws (T10) from the back of the module. 2. While pressing the release latch, pull the module casing slowly backwards and remove it from the main body. 3. Disconnect the front panel connector cable and the membrane keyboard cable from the Entropy board. 4. Detach the front cover (see 4.2.1). 5. Detach the Entropy board by removing the two screws located near the front chassis unit. Pull out the front chassis unit. 4.2.3 Reassembling the module Reverse the order of the disassembly steps. Check that: • • • screws are tightened properly cables are connected properly there are no loose objects inside the module 9-11 Document no. 2081505-001 B40/B20 Patient Monitor 5 Service parts 5.1 Ordering parts To order parts, Contact GE Healthcare. Contact information is available at www.gehealthcare.com. Make sure you have all necessary information at hand. NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout” after you have disassembled and reassembled the module. 5.2 Spare parts for E-ENTROPY Item Description Order No. 1 Module Casing, Single M1021035 2 SCREW, machine, M3x4mm, DIN7985, torx head, pan head, steel * 3 Entropy Board, E-ENTROPY M1093919 9-12 Document no. 2081505-001 Service parts Item Description Order No. 4 SCREW, screw for plastic, x12mm, WN1452, torx head, pan head, steel, zinc, SCREW-PT, PAN-HEAD, TORX, 3.0x12mm, ST-ZN, WN1452 * 5 SCREW, screw for plastic, x8mm, WN1452, torx head, pan head, steel, zinc, SCREW-PT, PAN-HEAD, TORX, 3.0x8mm, ST-ZN, WN1452 * 6 IMC CONNECTOR BOARD, M-ENTROPY 8004791 7 Front Chassis Unit, E-ENTROPY M1024285 8 Membrane Keyboard M1012126 9 SCREW, machine, M3x8mm, DIN965, torx head, flat countersunk head, steel * 10 Metal frame 879184 11 Ferrite holder, M-ENTROPY, Investment cast M1002842 12 CONNECTOR CABLE, M-ENTROPY..01 8005305 13 Torsion Spring M1020935 14 Latch M1021039 15 Front Cover, EN, E-ENTROPY M1021097 15 Front Cover, CS, E-ENTROPY M1063322 15 Front Cover, DA, E-ENTROPY M1024882 15 Front Cover, ES, E-ENTROPY M1024939 15 Front Cover, FI, E-ENTROPY M1024941 15 Front Cover, FR, E-ENTROPY M1024942 15 Front Cover, HU, E-ENTROPY M1046108 15 Front Cover, IT, E-ENTROPY M1024943 15 Front Cover, JA, E-ENTROPY M1024944 15 Front Cover, NL, E-ENTROPY M1024945 15 Front Cover, NO, E-ENTROPY M1024946 15 Front Cover, PL, E-ENTROPY M1024947 15 Front Cover, PT, E-ENTROPY M1024948 15 Front Cover, SV, E-ENTROPY M1024949 15 Front Cover, DE, E-ENTROPY M1020825 *) Part is not available from GE. Source locally. 9-13 Document no. 2081505-001 B40/B20 Patient Monitor 9-14 Document no. 2081505-001 ElectroMagnetic Compatibility APPENDIX A: ElectroMagnetic Compatibility Table 1 Guidance and manufacturer’s declaration – electromagnetic emissions Guidance and manufacturer’s declaration – electromagnetic emissions The monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance RF emissions CISPR 11 Group 1 The monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11 Class A Hermonic emissions Class A The monitor is suitable for use in all establishments other than domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. IEC 61000-3-2 Voltage fluctuations/ Class A flicker emissions IEC 61000-3-3 A-1 Document no. 2081505-001 B40/B20 Patient Monitor Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact ±6 kV contact ±8 kV air ±8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical fast transients/bursts IEC 61000-4-4 ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5 ±1 kV differential mode ±2 kV common mode ±1 kV differential mode ±2 kV common mode Mains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply lines IEC 61000-4-11 <5% UT (>95% dip in UT) for 0.5 cycle <5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 40% UT (60% dip in UT) for 5 cycles Mains power quality should be that of a typical commercial or hospital environment. If user of the equipment requires continued operation during power mains interruptions, it is recommended that the equipment be powered from an uninterruptible power supply or a battery. 70% UT (30% dip in UT) for 25 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 sec <5% UT (>95% dip in UT) for 5 sec 3 A/m 3 A/m Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 NOTE UT is the a.c. mains voltage prior to application of the test level. A-2 Document no. 2081505-001 Power frequency magnetic field should be at levels characteristic of a typical location in a typical commercial or hospital environment. ElectroMagnetic Compatibility Table 3 Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the equipment, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz 3 Vrms Radiated RF IEC 61000-4-3 3 V/m 80 MHz to 2.5 GHz 3 V/m 3.5 -----V1 P 3.5 -----E1 P 7 -----E1 P 80 MHz to 800 800 MHz to 2.5 where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. A-3 Document no. 2081505-001 B40/B20 Patient Monitor Table 4 Recommended separation distances between portable and mobile RF communications equipment and the monitor Recommended separation distances between portable and mobile RF communications equipment and the monitor. The monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the equipment as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 3.5 -----V1 P 80 MHz to 800 MHz 3.5 -----E1 P 800 MHz to 2.5 GHz 7 -----E1 P 0.01 0.12 0.12 0.23 0.1 0.37 0.37 0.74 1 1.17 1.17 2.33 10 3.69 3.69 7.38 100 11.67 11.67 23.33 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. A-4 Document no. 2081505-001 Installation and checkout form, B40/B20 APPENDIX B: Installation and checkout form, B40/B20 Customer Service Service engineer Date Monitor license MAC Address License key Modules Entropy module: Gas module: OK = Test OK Visual Inspection N.A. = Test not applicable Fail = Test failed Expected result Observed result OK N.A. Fail Expected result Observed result OK N.A. Fail 3.1. Visual inspection Functional Inspection 3.2.1. Start-up 3.2.2. Display 3.2.3. Time and date 3.2.4. Parameters measurements 3.2.5. Recorder 3.2.6. MC or S/5 Network connection 3.2.7. Conclusion B-1 Document no. 2081505-001 B40/B20 Patient Monitor Functional Inspection Notes Signature B-2 Document no. 2081505-001 Expected result Observed result OK N.A. Fail Maintenance and checkout form, B40/B20 APPENDIX C: Maintenance and checkout form, B40/B20 Customer Service Service engineer Date Monitor Installation IBP: SpO2 module: Gas module: Entropy module: Network: TEMP: Measuring equipment used: Equipment / tool: Manufacturer: OK = Test OK Visual Inspection Model/Type/Part Number: N.A. = Test not applicable Serial Number / ID: Calibration Date: Fail = Test failed Expected result Observed result OK N.A. Fail Expected result Observed result OK N.A. Fail 4.1. Visual inspection Notes Electrical Safety Tests 2.2. Power Outlet Test 2.3. Power cord and plug 2.4. Ground (earth) integrity Ground continuity test Impedance of protective earth connection < 0.1 ohms/ < 0.2 ohms < 0.1 ohms/ < 0.2 ohms C-1 Document no. 2081505-001 B40/B20 Patient Monitor Electrical Safety Tests Expected result Observed result OK N.A. Fail Observed result OK N.A. Fail 2.5. Earth leakage current test Normal Condition (NC) Single Fault Condition (SFC) < 500 µA < 300 µA < 1 mA 2.6. Enclosure (Touch) leakage current test Normal Condition (NC) Single Fault Condition (SFC) < 100 µA < 500 µA < 300 µA 2.7.1. Patient (source) leakage current test Normal Condition (NC) < 10 µA Single Fault Condition (SFC) < 50 µA 2.7.2. Patient (sink) leakage current tests < 50 µA Notes Functional Inspection 4.2.1. Start-up 4.2.2. Display 4.2.3. Keyboard(s) 4.2.4. Time and date Notes 4.2.5. Hemo Module . ECG measurement 1. Normal Sinus Rhythm 2. Pacemaker Detection 3. Asystole Detection 4. Leads Off Detection C-2 Document no. 2081505-001 Expected result Maintenance and checkout form, B40/B20 Functional Inspection Expected result Observed result OK N.A. Fail Notes . Respiration measurement 5. Respiration Rate 6. Apnea Detection Notes . Temperature measurement 7. Temperature detection Notes . Invasive blood pressure measurement 8. Zeroing 9. Static Pressure 10. Pressure Waveforms Notes . SpO2 measurement 11. Test measurement Notes . Non Invasive Blood Pressure measurement 12. NIBP Leak Test 13. NIBP calibration 14. NIBP hose detection Notes 4.2.6. Loudspeaker 4.2.7. Monitor software 4.2.8. Watchdog circuitry 4.2.9. Network C-3 Document no. 2081505-001 B40/B20 Patient Monitor Functional Inspection Expected result Observed result OK N.A. Fail Expected result Observed result OK N.A. Fail 4.2.10. Final cleaning Notes Monitor battery maintenance 4.3.3. Test the battery charge Notes Used Spare Parts Notes Signature C-4 Document no. 2081505-001 Appendix D, Service check form, E-miniC APPENDIX D Service check form, E-miniC Customer Monitor S/N Service Module type S/N Service engineer Date Measuring equipment / test gases used: Equipment / tool / gas: Manufacturer: Model/Type/Part No: Serial Number/ID: Cal Due Date: Prior to testing verify all equipment is calibrated via “Cal” labeling and record Cal Due Dates OK = Test OK Visual Inspection N.A. = Test not applicable Fail = Test failed Expected result Observed result OK N.A. Fail Expected result Observed result OK N.A. Fail 2.2. Visual inspections Notes Functional checkout 1. Gas sampling system leak test 2. Sample flow check 3. Zero valve operation 4. Gas calibration 5. Ambient pressure 6. Occlusion detection 7. Air leak detection 8. Airway gases 9. Apnea detection D-1 Document no. 2081505-001 B40/B20 Patient Monitor Functional checkout Notes Used spare parts Signature D-2 Document no. 2081505-001 Expected result Observed result OK N.A. Fail Appendix E, Service check form, E-sCO, E-sCAiO and N-CAiO APPENDIX E Service check form, E-sCO, E-sCAiO and N-CAiO Customer Monitor S/N Service Module type S/N Service engineer Date Measuring equipment / test gases used: Equipment / tool / gas: Manufacturer: Model/Type/Part No: Serial Number/ID: Cal Due Date: Prior to testing verify all equipment is calibrated via “Cal” labeling and record Cal Due Dates OK = Test OK Visual Inspection N.A. = Test not applicable Fail = Test failed Expected result Observed result OK N.A. Fail Expected result Observed result OK N.A. Fail 2.2. Visual inspections Notes Functional checkout 1. Gas Sampling System Leak Test 2. Sample Flow Rate Check 3. Reference Gas Flow Rate Check 4. Fan 5. Zero Valve Operation 6. Gas Calibration 7. Agent Identification 8. Ambient Pressure E-1 Document no. 2081505-001 B40/B20 Patient Monitor Functional checkout 9. Occlusion detection 10. Air Leak detection 11. Gas exhaust blockage 12. Airway Gases 13. Apnea detection Notes Used spare parts Signature E-2 Document no. 2081505-001 Expected result Observed result OK N.A. Fail Appendix F, Service check form, E-Entropy APPENDIX F Service check form, E-Entropy Customer Monitor S/N Service Module type S/N Service engineer Date Measuring equipment / test gases used: Equipment / tool / gas: Manufacturer: Model/Type/Part No: Serial Number/ID: Cal Due Date: Prior to testing verify all equipment is calibrated via “Cal” labeling and record Cal Due Dates OK = Test OK Visual Inspection N.A. = Test not applicable Fail = Test failed Expected Result Observed Result OK N.A. Fail Expected Result Observed Result OK N.A. Fail 2.1. Visual inspections Notes Electrical Safety Tests 2.7.1. Patient (source) leakage current test Normal Condition (NC) < 10 µA Single Fault Condition (SFC) < 50 µA 2.7.2. Patient (sink) leakage current tests using a test body < 10 µA using a patient cable < 50 µA Notes F-1 Document no. 2081505-001 B40/B20 Patient Monitor Functional Check 1. Module and sensor recognition 2. Sensor checking 3. Functional check with a simulator Notes Used spare parts Signature F-2 Document no. 2081505-001 Expected Result Observed Result OK N.A. Fail Appendix G, Networking disclosure to facilitate network risk managment APPENDIX G Networking disclosure to facilitate network risk managment Purpose and scope This disclosure is intended to satisfy the requirements of IEC 60601-1:2005 clause 14.13 and IEC/ISO 80001-1:2010 clause 3.5 for disclosure of network related specifications, requirements and residual risks in order to facilitate the responsible organization’s risk management activities (e.g. pursuant to 80001-1) for their networks incorporating the B40 monitor. Purpose for B40 monitor connection to network • Providing realtime patient data (i.e. parameters, waveforms and alarms) to compatible network devices such as central stations. • Remote configuration (patient admission, alarm settings, etc.) from compatible network devices. • Printing to a compatible network printer. Network interface technical specifications Connection Name Mission Critical (MC) network port Physical network connection type IEEE 802.3-1998 10/100BaseT Ethernet Speeds and duplex modes supported 100 Mbps half and full duplex Default IP address (from factory) IP address: 172.16.X.X Subnet Mask: 255.255.0.0 Gateway: 172.16.254.254 IP addressing IPv4 static Qos Support IP layer DSCP tagging Connection Name S/5 network port Physical network connection type IEEE 802.3-1998 10/100BaseT Ethernet Speeds and duplex modes supported 100 Mbps half and full duplex Address Virtual Plug ID Network information flows Flow Name Realtime patient data Network connection on device MC network port, S/5 network port Usage Type/Function/Purpose Waveforms, parameters, trends, alarm settings, admit/discharge patient Licensed/optional/required Licensed: Unity or S/5 G-1 Document no. 2081505-001 B40/B20 Patient Monitor Communication partner Device/IP address/network MC: CIC, Aware Gateway, CARESCAPE Gateway, Mobile Care Server S/5: iCentral Middle layer protocols MC: UDP; S/5: DRI Application layer protocol and encoding MC: Unity; S/5: DRI Ports MC: 2000, 7000, 7001 Traffic characterization and bandwidth requirements MC: 50 Kbps streaming data per patient view, 11 views maximum for wired network S/5: Variable Flow Name Ping Network connection on device MC network port Usage Type/Function/Purpose Network troubleshooting and check Licensed/optional/required Required Communication Partner Device/IP address/network PC, other medical devices Middle layer protocols ICMP Application layer protocol and encoding NA Ports NA Traffic characterization and bandwidth requirements On-demand, user initiated Flow Name HL7 Network connection on device MC network port Usage Type/Function/Purpose HL7 outbound to EMR system Licensed/optional/required Licensed Communication Partner Device/IP address/network HL7 Server Middle layer protocols TCP Application layer protocol and encoding HL7 Ports customer definded, default port is 6000 Traffic characterization and bandwidth requirements 720 bps, on demand Flow Name Printing Network connection on device S/5 network port Usage Type/Function/Purpose Print G-2 Document no. 2081505-001 Appendix G, Networking disclosure to facilitate network risk managment Licensed/optional/required Required Communication Partner Device/IP address/network Printer Middle layer protocols DRI Application layer protocol and encoding PCL Ports NA Traffic characterization and bandwidth requirements Max: 47 Kbps, Avg: 27 Kbps Flow Name Software Transfer Network connection on device MC network port Usage Type/Function/Purpose Performs software upgrades Licensed/optional/required Required Communication Partner Device/IP address/network Service-PC Middle layer protocols TCP Application layer protocol and encoding HTTP Ports 10001 Traffic characterization and bandwidth requirements 10 Mbps Required characteristics and configuration of network for support • The network must meet the specific requirements above for all traffic flows associated with the subset of features, use cases and workflows required by the responsible organization’s users. • In addition, the network must be “flat” (i.e. limited to a single IP broadcast domain). Potential risks to safety, effectiveness or security resulting from failure of IT network to provide the required Loss of network connectivity can result in the following hazardous situations: − Missed alarm at a remote viewing station (bedside or CIC). − Complete or partial loss or deterioration of remote monitoring of waveform and parameter data at remote viewing device. Product mitigations: − Low alarm volume is increased if network communication fails − Audio off, audio pause are interrupted if network communication fails − User is notified of network communication failure, message is displayed until user acknowledges it − User is notified if a duplicate IP address is detected − User is notified if a duplicate unit/bed name G-3 Document no. 2081505-001 B40/B20 Patient Monitor In addition to the hazardous situations identified above, connection of the B40 monitor to a network that includes other equipment could result in other unidentified risks to patients, operators or third parties. The responsible organization should identify, analyze, evaluate and control these risks on an ongoing basis including after changes to the network, which could introduce new risks and require additional analysis. G-4 Document no. 2081505-001 GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI 53223 USA Tel:+ 1 414 355 5000 1 800 558 5120 (US only) GE Medical Systems Information Technologies GmbH Munzingerstrasse 5 79111 Freiburg Germany Tel: + 49 761 45 43 - 0 GE Medical Systems Information Technologies, a General Electric Company, going to market as GE Healthcare www.gehealthcare.com 0459