7 Diagnostics and troubleshooting 7.1 Introduction This section contains diagnostic information including: • • • • • • 7.1.1 Instrument initialization EPROM-resident diagnostics PC software diagnostic folders Error log Windows Registry Information required when requesting assistance Important information You must use the methods described in this section when working with the diagnostic facilities provided with the Varian 700ES Series instrument. All of the following points should be read and understood to avoid the risk of damage to the instrument or abnormal operation. The DIP switches on the 80C32 controlled pwbs will alter the operation of the instrument. Before attempting to use the instrument for normal operation, ensure that all DIP switches are set to normal run position and the instrument power switch is turned off and on. 7.2 Instrument initialization This section provides information about instrument initialization routines which are started when the low power switch is turned on or when a system reset is generated. 7.2.1 386EX Control Board (02 101689 90) Application of low power to the 80C32 microprocessor on the 386EX control pwb produces an instrument system reset signal. This signal resets the gas control and RF interface pwbs. A system reset can be forced by pressing a switch on the 386EX control pwb. During the reset, the 80C32 samples the state of DIP switches SW2-2 to SW2-6, turns LEDs D20, D21, and D22 on and then performs a four second RAM test. When the 386EX control pwb has initialized, all LEDs are turned off, except for D20 which is flashed periodically. This indicates that the RS-422 serial communications link is working properly. Varian 700-ES Series instruments When the instrument has finished initializing the seven-segment LED will flash the firmware version to the first significant digit. 386EX control board: Normal startup sequence LED Duration Comment Decimal point only 1-2s Reset 1 < 1s Flash memory CRC test (boot block only). 2 < 1s RAM test (all of RAM). 3 2-3s Flash memory CRC tests (main instrument firmware). 4 2s All OK - waiting for PC-initiated flash download. (FWUpdate.exe) 6 3s Attempting to home the pre-optics mirror M1 X,Y (alternating) forever X, Y are the firmware version X.Y (eg. 1.0) X will display while the pre-optics is initialized, then the display will alternate X then Y to show version number. X will be on longer than Y. (Boot Block) 386EX control board: Startup sequence if FWUpdate.exe running LED Duration Comment Blank 1-2s Reset 1 < 1s Flash memory CRC test (boot block only). 2 < 1s RAM test (all of RAM). (Boot Block) Switch diagnostics 1, 2 and 11 (the normal boot sequence is interrupted at this point). 3 2-3s Flash memory CRC tests (main instrument firmware). 4 < 2s All OK - waiting for PC-initiated flash download. 7 indefinite In download mode, waiting for command from the PC, or returning version information to the PC. 8 < 60s Erasing flash memory and downloading new program. 9 < 3s CRC test on new program. 0 (flashing rapidly) until reset All OK indefinite Main program CRC is bad: program cannot execute. Special Cases 6 (instead of 4) Note: Following a successful download, the firmware will generate a reset and the normal startup sequence will begin. 7-2 Publication date: 10/16/2006 Varian 700-ES Series instruments Startup error codes: 386EX control board: Boot block error codes (same as satellite board codes) LED Duration Comment 0 (flashing rapidly) until reset OK – download completed 1 (flashing rapidly) until reset Boot block CRC failure 2 (flashing rapidly) until reset RAM test failure 5 (flashing rapidly) until reset Flash memory access failure 6 (flashing rapidly) until reset Incompatible F/W (boot block or main instrument) and H/W 7 (flashing rapidly) until reset Flash memory programming failure 8 (flashing rapidly) until reset Unable to execute instrument program (internal F/W fault) 7.2.2 External Satellite Board (Boot Block Programmer) The Boot Block Programmer is not part of the instrument. It is used only to program the Boot Block of the flash memory on the 386EX control Board - see Section 13 for full details. The sequence during initialization and the error codes are contained in the following charts. Satellite Board: Normal Programming Sequence LED (satellite board) Duration Comment - < 1s Reset 1 < 1s Flash memory CRC test (satellite board copy program only) 2 < 1s RAM test (all of RAM) 6 < 1s Flash memory CRC test (boot block image to be copied) 7 1-2s Erasing, programming and verifying boot block 0 (flashing rapidly) until reset All OK Publication date: 10/16/2006 7-3 Varian 700-ES Series instruments Satellite Board: Error Codes 7.2.3 LED (satellite board) Duration Comment 0 (flashing rapidly) until reset All OK - programming complete 1 (flashing rapidly) until reset Satellite board copy program CRC failure 2 (flashing rapidly) until reset RAM test failure 4 (flashing rapidly) until reset Boot block program image CRC failure 5 (flashing rapidly) until reset Flash memory access failure 6 (flashing rapidly) until reset Incompatible F/W (boot block or main instrument) and H/W 7 (flashing rapidly) until reset Flash memory programming failure Gas control pwb (02 101595 90) The gas control pwb is reset by the system reset signal from the 386EX control pwb. During a reset, the 80C32 samples the state of DIP switches SW1-2 to SW1-6, momentarily turns LEDs D4, D5, D6 and D7 on and does a four second RAM test. The 80C32 then reads the state of the argon inlet pressure switch. When the gas control pwb has initialized, all LEDs are turned off, except for D4 which is flashed periodically. This indicates that the RS-422 serial communications link is working properly. 7.2.4 RF interface pwb (02 101498 90) The RF interface pwb is reset by the system reset signal from the 386EX control pwb. During a reset, the 80C32 samples the state of DIP switches SW1-2 to SW1-6, momentarily turns LEDs D4 and D5 on and does a four second RAM test. IC9 is reset to ensure both the high voltage anode supply and filament supplies are disabled during initialization. When the RF interface pwb has initialized, D4 is flashed periodically. This indicates that the RS-422 serial communications link is working properly. 7.2.5 Peristaltic pump pwb (02 102029 90) The Peristaltic pump pwb is reset by the system reset signal from the 386EX control pwb via the Gas control pwb. A reset switch is also provided to reset the device. During a reset, the MCU samples the state of the jumpers to provide access to the various diagnostic modes. 7-4 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.3 EPROM diagnostics All four of the microprocessor-based pwbs (386EX control, gas control, RF interface and Peristaltic pump pwbs) are fitted with diagnostic switches/jumpers and LED’s. The diagnostic tests invoked by the diagnostic switches/jumpers test various functions of the instrument and should be used to help you find and fix faults. All switch driven, EPROM diagnostic tests will work when the pwb has been removed from the instrument for bench testing. All that is needed is the power supply connections to run the pwb. This feature allows you to run the full range of diagnostic tests while bench-testing the pwbs. To run any of the diagnostic tests, select the required switch combination and enable the reset function (turn on the low voltage). Apply +5 VDC to the 80C32 microprocessor (if bench testing), or create an instrument system reset at the 386EX control pwb to start the test. The tests will continue to cycle until the switches are returned to the normal operating position and the pwb is re-intialized by a system reset. 7.3.1 386EX control pwb DIP switch settings are defined as: • • Open Closed = logic 1 = logic 0 = High = O = Low = CL Switch SW1 generates a system reset on the 386EX control board which resets all boards. SW3 is a rotary switch which is normally set to zero and is reserved for future use. DIP switch SW2 on the 386EX control pwb performs two tests: SW2 6 5 4 3 2 1 X CL CL CL CL O Test X CL CL CL CL O/CL Note Publication date: 10/16/2006 Operation Normal operation IEEE communications time out Check that SW1-1 on the gas control pwb is set to closed (normal operating mode) upon the completion of all tests. 7-5 Varian 700-ES Series instruments Communications Test The 386EX control pwb and PC software communicate by the IEEE RS 488 parallel link. With SW2-1 set to open, the plasma is automatically extinguished if the PC fails to communicate with the instrument for a period of more than ten seconds. Set SW2-1 closed before power up to disable this feature. Setting SW2-1 closed will allow the plasma to stay on if the PC does not communicate with the instrument for extended periods of time. Note Ensure that switch SW2-1 is set open when you have finished testing. Camera PWB simulator The 386EX control pwb includes circuitry that simulates the camera assembly. The simulator has been designed as a standalone test circuit that simulates camera pwb data and control functions. The simulator tests the control board, including the plug, the interconnecting cable, and the PC interface. To connect and run the camera simulator: 1. Turn off the instrument and remove the covers from the right side of the instrument to allow access to the camera and 386EX control board. 2. Disconnect the cable connecting the 386EX control board to camera board at the camera board end. 3. Plug the now disconnected cable into the test plug, PL4, on the 386EX control board. You may need to swing the 386EX control board out to gain access. 4. Turn on the instrument. 5. Check LED’s D13 and D15 to D19 are lit. They indicate relevant signals are passing through the camera cable and can be used as a quick check of the cable and supplies. 6. Run the Varian 700-ES Series instrument PolyAlign software and select the Location folder. 7. Press the “Read spectrum” button and view the echellogram. 8. If everything is working properly you should see a ramp across both the left and right side of each DLA for the Varian 720/730-ES Series instruments, or a set of nested diamonds for the Varian 710-ES Series instrument as shown below. 7-6 Publication date: 10/16/2006 Varian 700-ES Series instruments Varian 720/730-ES series loopback echellogram Varian 710-ES series loopback echellogram Publication date: 10/16/2006 7-7 Varian 700-ES Series instruments 7.3.2 Gas control pwb The diagnostic tests are selected by the DIP switch SW1. O = switch open CL = switch closed X= switch open or closed (don’t care) Test * 1 2 3 4 5 6 7 8 9 10 11 12 1 CL O CL CL CL CL CL CL CL CL CL CL CL SW1 Switch Positions 2 3 4 5 O O O O O O O O O O O CL O O O CL O CL CL CL O O CL CL O CL CL CL O O CL CL O O CL CL O O CL CL O O CL CL O O O CL O O CL CL Test Description 6 X X X X X X X X X X X X X System reset disable (normal operation) System reset enable LED D4-D7 Gray code test 8031 serial port test SV1 to SV7 & igniter simultaneous SV1 to SV6 sequence SV9-10-8-11 sequence Peristaltic pump at 25RPM & MFC at 1SLPM RAM test 8031 Ports 1&3 IC14 input latch - HI nibble IC14 input latch - LO nibble Peristaltic pump & MFC ramp Test 1 - System reset enable Allows the Gas Control pwb to be reset externally, via OP1, from the 386EX control pwb. Test 2 - LED D4-D7 Gray code test Cycles LEDs D4 to D7 in a Gray code pattern Test 3 - 8032 serial port test IC7 outputs a pulse train on pin 11 (TXD) Test 4 - SV1 to SV7 & igniter simultaneous Simultaneously activates the solenoid valves controlling plasma and auxiliary gas flows (SV1 to SV7) and the igniter. LEDs D4 to D7 on the Gas Control pwb, and LEDs D1 to D8 on the Gas Box Interconnect pwb, are ON when the solenoid valves and igniter are ON. Test 5 - SV1 to SV6 sequence Activates the solenoid valves SV1 to SV6 in sequence. Each solenoid is activated for around 1 second, at which time the corresponding LED (D1 to D6) on the Gas Box Interconnect pwb, is ON. LEDs D4 to D7 on the Gas Control pwb are ON for 6 seconds, OFF for 1 second. 7-8 Publication date: 10/16/2006 Varian 700-ES Series instruments Test 6 - SV9-10-8-11 sequence Activates solenoid valves SV9-SV10-SV8-SV11, in sequence. Each solenoid is activated for around 1 second, at which time the corresponding LED (D9 to D12) on the Gas Box Interconnect pwb is ON. LEDs D4 to D7 on the Gas Control pwb are ON for 3 seconds, OFF for 1 second. Test 7 - Peristaltic pump at 25RPM & MFC at 1SLPM Sets the peristaltic pump speed to 25rpm (TP13=2.45V) and the MFC flow rate to 1SPLM (TP18=3.3V). LED D5, D6 & D7 reflect MFC status. LED D4 ON LED D5 ON LED D6 ON LED D7 ON pwb is 02.10595.00 MFC not connected MFC HIGH FLOW (TP19 ≥ 4V) MFC LOW FLOW (TP19 ≤ 2.7V) Test 8 - RAM test Performs a simple test of the RAM, IC12. All LEDs OFF indicates that the RAM is OK Test 9 - 8031 Ports 1&3 Writes a series of test value to ports P1 (P1.0 to P1.7) and P3 (P3.0 to P3.5) of IC7. Data sequence is 00, 55, AA, FF at 1/2 second intervals. Note that the LEDs D4 to D7 are driven by port P1. Test 10 - IC14 input latch - HI nibble Continuously reads input latch IC14 and displays the high four bits on LEDs D4-D7. A high logic level at the input of IC14 turns the corresponding LED ON. LED D4 ON LED D5 ON LED D6 ON LED D7 ON Publication date: 10/16/2006 IC14 pin6 HI IC14 pin7 HI IC14 pin8 HI IC14 pin9 HI LOW argon pressure LOW nebuliser pressure MFC HIGH FLOW MFC LOW FLOW 7-9 Varian 700-ES Series instruments Test 11 - IC14 input latch - LO nibble Continuously reads input latch IC14 and displays the low four bits on the LEDs D4-D7. A high logic level at the input of IC14 turns the corresponding LED ON. LED D4 ON LED D5 ON LED D6 ON LED D7 ON IC14 Pin2 HI IC14 Pin3 HI IC14 Pin4 HI IC14 Pin5 HI MFC not connected VAC sense (PL4 Pin6) ? VAC option (PL4 Pin1) Test 12 - Peristaltic pump & MFC ramp The peristaltic pump slowly ramps down from full speed to zero speed, as TP13 ramps down from 4.9V to 0V (10 seconds). The MFC slowly ramps up from 0 SLPM to 1.5 SLPM, as TP18 ramps up from 0V to 5.0V (in 10 seconds). 7.3.3 RF interface pwb There are three functions assigned to DIP SW1 on the RF interface pwb: 0= Open, 1= Closed SW1 Test 1 Test 2 Test 3 123456 000000 111111 101111 110111 Operation Normal operation Local reset LED test 80C32 serial port test RF Interface Test 2 This test is included in all microprocessor-driven pwbs. When it is enabled, the test cycles LEDs D3, D4, D5 and D6 in a Gray code pattern. RF Interface Test 3 This test is included in all microprocessor driven pwbs. When it is enabled, the 80C32 outputs a pulse train at the TXD output. Set DIP switch 3 closed position and use an oscilloscope to sample TXD on pin 11 of IC1. 7-10 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.3.4 Peristaltic pump pwb There is a major and a minor set of diagnostic modes for the Peristaltic pump pwb. The major diagnostic mode is sensed from the jumper settings immediately after hard reset (or power up) and locked in until the next reset (or power down). The minor diagnostic mode is continuously sensed and acted on in the context of the major diagnostic mode. During diagnostics, the LEDs may be used differently from normal operation as described below. Major Mode Minor Mode Mode Jumper 2 Jumper 1 Jumper 2 Jumper 1 open open open open 00 open open open inserted 01 open open inserted open 02 open open inserted inserted 03 open inserted x x 1X inserted open open open 20 inserted open open inserted 21 inserted open inserted open 22 inserted open inserted inserted 23 inserted inserted inserted inserted inserted inserted open open inserted open inserted open 30 31 32 inserted inserted inserted inserted 33 Description Normal Operation. Normal Indications. Normal Operation. Supply Voltage (3 digits) shown on red LED. Normal Operation. Control Frequency (up to 3 digits) shown on red LED. Normal Operation. Normal Indications. not implemented, don’t use Motor Drivers off, PWM operating 60sec @ 15 RPM, 40sec @ 50RPM 40sec @ 15 RPM, 60sec @ 50RPM 4sec @ 15 RPM, 6sec @ 50RPM Motor Speed = 15rpm Motor Speed = 50rpm Motor Speed = 5rpm Motor Drivers off, PWM operating Operating Indicators LED’s on the Peristaltic pump pwb are provided for diagnostic purposes. The jumpers share the IO pins with the LEDs. The jumpers are scanned in turn and monitored to detect the presence of a jumper. Publication date: 10/16/2006 7-11 Varian 700-ES Series instruments Below is a description of LED’s during normal operation (not in diagnostics mode). Red LED After reset, the red LED flashes the Firmware Version once and then remains off. If the red LED keeps flashing an error has been detected and the flashing represents a 2 digit error code which is cycled continuously. Yellow LED Intensity: With the frequency input below the turn-on threshold, the LED is dimly illuminated (25% duty cycle) – this indicates that the motor is off and the µC's timer function works properly. With the frequency input above the turn-on threshold, the LED is brightly lit and the motor is energized and running. Flicker: If the LED shows a single flicker, approx once per second, then the firmware has detected the small motor (3-channel pump). If the Led shows a double flicker, approx once per second, then the firmware has detected the large motor (4-channel pump). The single or double flicker is present regardless of the intensity. Green LED This is the Water Flow Indicator. The actual flow detection and reporting is done in hardware without any MCU involvement. The MCU merely indicates the state of the Water Flow through the green LED. No water flow detected: If the LED flashes briefly once every 2 seconds (approx) then no water flow has been detected and the brief flash indicates that the LED and µC function correctly. Water flow adequate: If the LED flickers quickly and brightly then water flow has been detected. In this state, the flicker indicates that the LED and MCU function correctly. The flicker rate has no relationship to the rate of water flow – the flicker rate is fixed in firmware. Serial Port If Jumper 3 was present at reset, then the Firmware Version, Motor detection ADC counts, calculated Motor Inductance values and subsequent motor selection (“M1” or “M2”) are sent to the serial port. Serial port settings: 1 start, 8 data, 1 stop bits, no parity, 38400 bps. 7-12 Publication date: 10/16/2006 Varian 700-ES Series instruments Error Codes Identifier Code Description ERROR_PB2_STUCK_LOW 12 PUMP_EN stuck in logic low ERROR_PB2_STUCK_HIGH 13 PUMP_EN stuck in logic high ERROR_PD3_STUCK_LOW 14 phase-2 (cosine) stuck in logic low ERROR_PD3_STUCK_HIGH 15 phase-2 (cosine) stuck in logic high ERROR_PD5_STUCK_LOW 16 phase-1 (sine) stuck in logic low ERROR_PD5_STUCK_HIGH 17 phase-1 (sine) stuck in logic high ERROR_PD6_STUCK_LOW 18 phase-3 stuck in logic low ERROR_PD6_STUCK_HIGH 19 phase-3 stuck in logic high ERROR20_NO_MOTOR 20 no motor detected ERROR21_PHASE1_OPEN_CIRCUIT 21 no current in winding 1 detected ERROR22_PHASE2_OPEN_CIRCUIT 22 no current in winding 2 detected ERROR23_PHASE3_OPEN_CIRCUIT 23 no current in winding 3 detected ERROR25_PHASE_IMBALANCE 25 current difference exceeds 50% of current average ERROR26_PHASE1_SHORT_CIRCUIT 26 excessive current in winding 1 detected ERROR27_PHASE2_SHORT_CIRCUIT 27 excessive current in winding 2 detected ERROR28_PHASE3_SHORT_CIRCUIT 28 excessive current in winding 3 detected Morse Code The Firmware Version, error codes and some other numeric indications are given in international Morse code. This minimises the number of flashes that need to be counted, allows for multi digit indications and is quite easy to learn. Morse code digits always consist of 5 flashes per digit. There are short and long flashes, the pause between flashes within a digit are the same as the duration of a short flash. The pause between digits is longer. Hint: It is only necessary to either count the short or the long flashes and to remember if the short flashes were before or after the long ones. Morse Code Digit • 1 • • • • • 2 • • • • • • • • • • • 4 Example: Error Code “17” would be idicated in Morse code like this: • 5 • • • 6 • • 7 • • 8 • 9 3 1 7 0 Publication date: 10/16/2006 7-13 Varian 700-ES Series instruments 7.4 PC software diagnostics This section contains information on the diagnostic facilities provided in the Varian 700-ES Series instrument software. The software diagnostics are grouped as: • • • General access folders - the same as Instrument setup General diagnostics folders Service-only diagnostic folders The general access folders are common with the Instrument Setup page, and all users have access to these folders from the Instrument setup page. They include the Status folder, Details folder, Wavelength Calibration folder, Ignition folder, Echelle folder, Torch Align folder Autosampler folder and Log folder. No password is required for these folders. These folders have been made available to all users to aid in instrument fault diagnosis over the phone. 7-14 Publication date: 10/16/2006 Varian 700-ES Series instruments The general diagnostic folders include all the general access folders, which are not password-protected, and three additional folders, Quick folder, Gasbox folder and Optics folder, These folders are password-protected. They are accessible from the Instrument Setup page. To enable them, click on the flashlight icon and enter the password wallaby. Publication date: 10/16/2006 7-15 Varian 700-ES Series instruments The Service-only diagnostic folder (or extended diagnostics) includes all the general access folders, which are not passwordprotected, and the three additional general diagnostics folders as well as the RF folder. The RF folder is further password-protected. To enable this folder, click on the key icon in the top right corner of the diagnostics screen and enter the password virgil. 7-16 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.4.1 Diagnostics toolbar The diagnostics page has a toolbar with various functions: 1 2 3 4 5 6 7 8 9 Functions: 1. Continuous refresh. The status of the instrument is cyclically updated. Many parameters will be greyed out and inaccessible when in this mode. 2. Single update. The status of the instrument is updated once when you click on this icon. 3. Apply. The instrument is updated with any changed parameters when you click on this icon. Any time you change parameters on the diagnostics folders you must click on this icon to implement the change. 4. Torch On or Torch Off. These icons turn the torch on and off. 5. Pump On, Pump Off, Pump fast. These icons turn the peristaltic pump on or off, and the fast pump icon makes the pump run at 50 rpm. 6. Gasbox Power Down. This removes power to all solenoids, and disables the igniter and peristaltic pump. 7. RF Power Down. This removes the RF drive to the power tube, filament and high voltage supplies. 8. ICB, GB and RF “LEDS”. These icons are indicators showing when each of the three boards are communicating. The boards are the 386EX control Board, the Gasbox board, and the RF interface board. 9. Theoretical wavelength positions. This icons turns the wavelength calibration offsets on (button “depressed”) or off (button “out”). 10. The “key”. This is the Password entry icon. Entering the correct password allows you access to the extended service-only diagnostics. Publication date: 10/16/2006 7-17 10 Varian 700-ES Series instruments 7.4.2 Status folder The Status folder is the first folder displayed when entering the service diagnostics. Depending on the mode of entry (general or extended) up to twelve folders are displayed. The Status folder displays: 1. The firmware revision of all EPROMs and flash memory in the instrument. 2. Plasma status 3. RF, high voltage and filament status 4. Polychromator temperatures and purge status 5. Argon pressure switch status and flows 6. Peristaltic pump rate 7. Water flow status 8. Argon filter usage Note This information is available only if the PC is connected to the instrument and the IEEE communication is working. Status folder 7-18 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.4.3 Details folder The Details folder shows serial numbers, installation and maintenance dates, and hours of running of the filament, shutter and filters. The software analyses the contents of the EEPROM on the 386EX Control Board and reports the relevant values. To change any of the fields you will need to enter the service password. Details folder 7.4.4 Wavelength Calibration folder The Wavelength Calibration folder shows the results from the last successful wavelength calibration. It also allows reading and storage of the Dark Current, and calculation of the electrical offset. Additionally, for Varian 710-ES Series instruments only, this folder also allows you to calibrate the shutter and the attenuator. For the Varian 720/730-ES Series instrument Publication date: 10/16/2006 7-19 Varian 700-ES Series instruments Dark current scan Click here to initiate a “read spectrum” of the camera to quantify the dark current characteristics. You must do this before performing a wavelength calibration. The results are stored and subtracted from all subsequent readings. This will also take readings and average them to produce an average offset of the camera. Calibrate Click here to initiate a wavelength calibration. The results will be updated when the spectrum has finished. The results will also be appended to a file called “Wavelength Calibration.txt” in the C:\Program Files\Varian\ICPExpert II 7XX-ES\Run\Log subdirectory. Calculate drift This action will calculate the drift since the last calibration and display the result. Excessive drift may suggest recalibration. Stop (F12) This action will stop any of the actions above. Store Dark Current This exports the results of the dark current scan to a text file which is tab delimited, suitable for importing into a spreadsheet. This file contains a value for every pixel on the camera. This is the only file which contains data for every pixel. You have the option for naming this file when you save it, but it does not prompt you to overwrite if you have a previously saved file with the same name. For the Varian 710-ES Series instruments 7-20 Publication date: 10/16/2006 Varian 700-ES Series instruments Calibrate Detector Click here to initiate a “read spectrum” of the camera to quantify the dark current characteristics. You must do this before performing a wavelength calibration. The results are stored and subtracted from all subsequent readings. This will also take readings and average them to produce an average offset of the camera. Calibrate Wavelength Click here to initiate a wavelength calibration. The results will be updated when the spectrum has finished. The results will also be appended to a file called “wavelengthcalibration.txt” in the C:\Program Files\Varian\ICPExpert II 71X-ES\Run\Log subdirectory. Calculate drift This action will calculate the drift since the last calibration and display the result. Excessive drift may suggest recalibration. Calibrate shutter This will cause the instrument to take photometric measurements which characterize the camera exposure time errors caused by the mechanical inertia and reproducibility variation of the shutter. Typically, the shutter delay will be in the range ±2 mS. Results of this test can be found in the ShutterCalibration.txt file located in the C:\Program Files\Varian\ICPExpert II 7XX-ES\Run\Log subdirectory. Calibrate Attenuator The attenuator calibration task is designed to characterize the light transmission of the attenuator across the wavelength range of the instrument. It will generate a pass/fail message as compared against expected values. The results of this test can be found in the AttenuatorCalibration.txt file located in the C:\Program Files\Varian\ICPExpert II 7XX-ES\Run\Log subdirectory. Note Publication date: 10/16/2006 The Calibrate Shutter and Calibrate Attenuator options are available independently if the service diagnostics password has been entered. If the service password is not entered, these two options are combined under a single option called Calibrate Hardware. 7-21 Varian 700-ES Series instruments Store Dark Current This exports the results of the dark current scan to a text file which is tab delimited, suitable for importing into a spreadsheet. This file contains a value for every pixel on the camera. This is the only file which contains data for every pixel. You have the option for naming this file when you save it, but it does not prompt you to avoid overwriting if you have a previously saved file with the same name. 7.4.5 Ignition folder This folder shows the parameters for plasma ignition. The un-shaded parameters may be modified; however, the parameters chosen (as defaults) are the best compromise for both axial and radial ignition under normal operating conditions. The shaded parameters are not adjustable. They can only be altered in the Windows Registry. 7-22 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.4.6 Torch Alignment The torch alignment folder allows you to set the position of the pre-optics scanning mirror (M1) so that it is looking at the optimum part of the plasma. In an axial instrument there are two settings - horizontal and vertical - while on a radial instrument there is only the horizontal setting (the vertical setting - called viewing height - is stored in the worksheet). Note For Radial instrument, if the service password has been entered, the vertical setting can also be monitored. To perform the alignment you aspirate a 5 ppm Mn solution and perform a horizontal and then a vertical torch scan (if it is an axial instrument), using the supplied torch alignment worksheet. Unless the vertical scan is very close to the home position, you should do another vertical scan. When the instrument scans the torch, it sets the pre-optics to the optimum position for viewing the plasma. Publication date: 10/16/2006 7-23 Varian 700-ES Series instruments 7.4.7 Autosampler (SPS) folder The Autosampler folder allows you to control the SPS3 autosampler. You can: • Align the probe • Perform a rinse • Set the probe height • Send the probe to any sample or standard position, simply by clicking on the visual position. • Park the probe, either to the left or the right. You can change the Rack Type by Right Clicking on any rack position and selecting the appropriate rack. 7-24 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.4.8 Log folder The Varian 700-ES Series instrument software records all error messages that occur when operating the instrument. The errors in the log can be viewed using this folder. The errors are also recorded in a file called error.log and stored in the C:\Program Files\Varian\ICPExpert II 7XX-ES\Run\Log subdirectory. 7.4.9 Graphs Although this is not a separate folder, graphs form part of the next four diagnostics folders. The same graph information and stored files are displayed on these folders: • Quick • Gasbox • RF The graphs (spectrums) are visual representation of signals from the camera, and there are several varieties on the types of graphs available. Individual or multiple DLA responses, either the right or left side, can be displayed and they can be read repetitively or singly. Collections of pre-defined spectrums can be stored and retrieved from the registry, using the “Save” and “Load” buttons. Note Publication date: 10/16/2006 For the Varian 720/730-ES series only. If you position the cursor over the “Save” button and hold down the <Ctrl> and <Shift> keys together, with at least one graph in the list, you can generate a list of graphs which covers every DLA, and it will be saved to the registry as “all”. 7-25 Varian 700-ES Series instruments Graph of 6 DLAs using an echellogram for the Varian 720/730ES series instrument. 7.4.10 Graphs Toolbar The graphs section has its own toolbar with various functions: 1 5 2 3 4 8 7 6 9 10 Functions: 7-26 1. Read Spectrum. This initiates a single read of the specified graph(s). 2. Read Continuous Spectrum. This initiates a cyclic, or continuous, read of the specified graph(s). 3. Stop. This icon becomes active (red) once a read spectrum has begun. It STOPS the read. 4. New Graph. This opens a new graph and pops up the graph parameters pages: Publication date: 10/16/2006 Varian 700-ES Series instruments On these pages you can choose to specify the DLA, left or right side, or the element and wavelength. You can also specify the integration time to increase or decrease the “gain”. Instead of a wavelength read you can specify a time scan. 5. Edit. This icon allows you to edit the graph parameters. 6. Delete. This deletes the specified spectrum. 7. Move. These will move the focus up or down the list of spectra. 8. Multiple Graphs. Selecting this icon will display the spectrum as a series of graphs (instead of an echellogram): 9. Single Graph. Selecting this icon will display each graph singly. 10. Echelle. Selecting this icon will display the camera’s response as an echellogram. Publication date: 10/16/2006 7-27 Varian 700-ES Series instruments 7.4.11 Quick Folder The Quick folder is a collection of the most commonly accessed items from all the other diagnostics folders. It is probably the folder that you will use the most in a diagnostics session. 7.4.12 Gasbox folder The Gasbox folder allows testing, monitoring and control of the gas control unit including the igniter. A tick in the checkbox indicates selection. Gasbox folder 7-28 Publication date: 10/16/2006 Varian 700-ES Series instruments Plasma flow The plasma flow field allows plasma gas flow to be turned ON within allowable limits. The field also confirms selection. As the smallest increment of plasma flow is 1.5 L/min, all settings are rounded up or down to the nearest multiple of 1.5 L/min. Auxiliary flow The auxiliary flow field allows the auxiliary gas flow to be turned ON within allowable limits. The field confirms selection. The smallest increment of auxiliary flow is 0.75 L/min, all settings are rounded up or down to the nearest multiple of 0.75 L/min. Pump Rate The pump rate field allows the peristaltic pump to be set to any allowable rpm between 0 and 50 rpm. The field then confirms the selection. Auxiliary boost The auxiliary boost field allows auxiliary gas flow to be turned on or off. The ‘&‘ in the field confirms selection. The auxiliary boost flow is rated at 26 L/min. Snout/Cone purge The snout/cone purge field allows the snout/cone purge to be turned on or off. The ‘&‘ in the field then confirms selection. It cannot be turned off while the plasma is on. Poly purge (boost) The poly purge field allows the poly boost purge solenoid to be turned ON or OFF. The field confirms selection of ON by the use of a ‘&‘. Nebulizer (instruments fitted with a pressure regulator) The nebulizer field allows the nebulizer solenoid to be turned ON or OFF. The field confirms the selection of ON by the use of a ‘&‘. MFC flow (instruments fitted with a mass flow controller) The MFC flow field allows the nebulizer flow to be set to any allowable flow rate between 0 and 1.5 L/min. The field then confirms the selection. Publication date: 10/16/2006 7-29 Varian 700-ES Series instruments Argon pressure OK This is a read only field which displays the current state of the argon inlet pressure switch. ‘&‘ symbolizes closed (pressure OK) and o symbolizes open (pressure low). Igniter The igniter field allows the igniter to be turned ON to continuously generate a spark. The field confirms the selection of ON by the use of ‘&‘. To allow the spark to discharge through the torch, set the plasma flow greater than 15 L/min. Warning 7.4.13 Contact with the high voltage spark from the igniter can cause electric shock. To avoid electric shock, always check that the torch compartment door is closed when the igniter module is operating. Optics folder This folder allows control of the Pre-optics, temperature and polychromator. Pre-optics You can drive the pre-optics stepper motors independently to arbitrary positions or to the home sensor. A tick, ‘&‘, will appear in At Home check box when the opto sensor detects the mirror is “home”. 7-30 Publication date: 10/16/2006 Varian 700-ES Series instruments Temperature Control The Peltier cooling on the camera and the heater can be turned on by entering a tick, ‘&‘, in the adjacent box. You must click on the “apply” button on the toolbar to effect the change. The Peltier and casting temperatures can also be monitored. Polychromator Camera supplies supplies on. - A tick, ‘&‘, in this box will turn the +/- 15V Shutter open - A tick, ‘&‘, in this box will open the shutter, and a blank will close the shutters. Full illumination LED - A tick, ‘&‘, in this box will turn on the LED which floods the detector with light. 7.4.14 RF folder The RF folder allows testing, monitoring and control of the RF generating system. This folder is accessible only in extended service diagnostics (password protected). RF interlocks These are all read only fields. When the monitor column is ‘&‘, the interlock is being actively monitored. During normal plasma operation all of the interlock states should be ‘&‘. RF status lines These are all read only fields. The status lines reflect the current condition of each function. As each is enabled, the monitor field will display a ‘&‘. Publication date: 10/16/2006 7-31 Varian 700-ES Series instruments Filament (stage 1) The Set field allows the filament supply to be turned ON ‘&‘ without application of the high voltage anode supply to the power tube. The field confirms selection. HV (stage 2) The Set field allows the high voltage anode supply to be applied to the power tube. The ‘&‘ field confirms selection. Note The software will not allow the HV supply to be turned ON unless the filament supply (stage 1) is ON and has warmed the filament of the power tube for a minimum of 30 seconds. RF power The Set field allows the level of RF power produced by the RF generator to be changed. The Actual field confirms selection. Note This field allows the RF power output to be set to a maximum of 2.0 kW (keyboard) which is higher than allowed in normal operating mode (1.5kW – axial and 1.7 kW – radial). Filament turn-on time This is a read only field which logs the time the filament supply was turned-on. This value is stored in the EEPROM on the 386EX control board. Filament Low Voltage There are two voltages that the filament uses, 6.3 volts RMS and 5.9 volts RMS. After the filament has accumulated 200 hours of operation, the filament voltage drops from 6.3 volts to 5.9 volts. This box is an indication of which filament voltage is being used. If the box is ticked, ‘&‘, this indicates it is running on the lower voltage, 5.9 volts RMS. 7-32 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.5 Polychromator Align software There is a special function software program provided called PolyAlign.exe (polychromator alignment). This software gives the CSR direct feedback of the camera response to allow accurate alignment of the polychromator and camera. The software is available in the C:\Program Files\Varian\ICPExpert II 7XX-ES\Run subdirectory. To access this software, double click on the file called PolyAlign.exe in the run subdirectory or select Polychromator Align from the START>PROGRAMS>ICP EXPERT II menu. The Polychromator Align Password pop-up will appear. Enter the service password ‘virgil’. Note Publication date: 10/16/2006 The software displays the same information in a different way for the Varian 710-ES Series instrument as it does for the Varian 720/730-ES Series instrument. This is because of the different camera geometry. 7-33 Varian 700-ES Series instruments 7.5.1 Location folder The location folder allows you to view the echellogram in different areas and to locate specific peaks. Use this folder to perform DLA and wavelength screw adjustments on the focussing mirror see Optics chapter. Location folder for the Varian 720/730-ES Series instruments Location folder for the Varian 710-ES Series instruments 7-34 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.5.2 Set Focus folder The Set Focus folder concentrates on the Argon 617 and Argon 594 nm doublet and resolution of the Zinc 213.856 peak. Use this folder to adjust the focus of the polychromator using the three screws on the focussing mirror. Set Focus folder for the Varian 720/730-ES Series instruments Set Focus folder for the Varian 710-ES Series instruments Publication date: 10/16/2006 7-35 Varian 700-ES Series instruments 7.5.3 Focus Check folder The Focus Check folder shows a spectrum of lines across the CCD to check that the focus across the whole detector meets specification. Focus Check folder for the Varian 720/730-ES Series instruments Focus Check folder for the Varian 710-ES Series instruments 7-36 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.5.4 Rotation folder The Rotation folder displays the ratio of two argon lines on DLA 21 that span three DLAs. Both lines should have the same ratio when the CCD is aligned correctly. DLA 20 DLA 21 DLA 20 Ratio of DLA 20 and DLA 22 Rotation folder for the Varian 720/730-ES Series instruments Rotation folder for the Varian 710-ES Series instruments Publication date: 10/16/2006 7-37 Varian 700-ES Series instruments 7.5.5 Registration folder The Registration folder shows the intensity of a set of elements alongside the Argon ratio. This page is used to establish the optimum setting for the Argon ratio. Registration folder for the Varian 720/730-ES Series instruments 7-38 Publication date: 10/16/2006 Varian 700-ES Series instruments 7.6 Error log and Debug log The Varian 700-ES Series instrument software reports all the operational and user errors in the Operations log contained in each worksheet. These can be viewed using the Varian 700-ES Series instrument software. The Varian 700-ES Series instrument software records all error messages that occur when operating the instrument, in a text file called ‘error.log’. The error log is a file in the C:\Program Files\Varian\ICPExpert II 7XX-ES\Run\log subdirectory. It can be viewed and printed using any text editor. When the Varian 700-ES Series instrument software starts it renames the existing ERROR.LOG as ERROR.~LO and begins a new ERROR.LOG. This means that the old error log, ERROR.~LO is continually overwritten. To keep this error log you will need to rename it before starting the Varian 700-ES Series instrument software. Varian 700-ES Series instrument error messages are classified in three ways: • • • Warnings Terminal Fatal (W) (T) (F) The error log automatically traps and records only fatal messages which are flagged in the error log with ‘F’. Each record in the error log file is date and time stamped. The error log date is updated only when the Varian 700-ES Series instrument software is started. There is another log file in the same \log subdirectory, the DEBUG.LOG. This file is appended to constantly and just keeps getting bigger. However, the information in the debug file is only useful to software engineers and is not readable without the appropriate software knowledge. It may be needed in some circumstances by Varian software engineers to help diagnose problems. Publication date: 10/16/2006 7-39 Varian 700-ES Series instruments 7.7 Windows Registry Warning Before you consider altering anything in the Windows Registry you must backup the registry files. Alterations to the registry can cause catastrophic failure to the PC’s operating system and software. The simplest way to backup the registry is to “export” the registry as a registry file using Windows Registry Editor. It can be later “imported” using the same program. See below for a way to restore the registry if you do not have a back up. The Varian 700-ES Series instrument software uses the Windows registry extensively for setting and storing various functions. Many settings within the registry are integral to the Varian 700ES Series instrument software and should not be changed. Incorrect settings could cause the software or the Windows operating system to malfunction. There are some “User settings” which can be altered under direction from people experienced with the registry settings. The majority of those settings are contained in the registry under the tree: HKEY_LOCAL_MACHINE\Software\Varian\ICP\ES\ICP-OES 7XX Within the registry, the variable settings are referred to as “keys”, and they can be changed by double-clicking with the left mouse button. To make changes in the registry you must run the registry editor program, Regedit. To do this select “Run” from the start menu and enter “regedit” in the entry box: If you need to restore the default settings for Varian 700-ES Series instrument software in the registry, you must re-install the software, or “import” a valid registry file. 7-40 Publication date: 10/16/2006 Varian 700-ES Series instruments Restoring the Registry If the registry becomes corrupted you can restore it from Windows automatically stored backup from the last successful restart: 1. Start or Restart your computer. After your computer starts, Windows displays the following text: Starting Windows 2. While the text is on your screen, press and release the F8 key. A menu appears. 3. Choose the Command Prompt Only option. 4. At the command prompt, change to your Windows directory. For example, if your Windows directory is C:\Windows, you would type the following: cd c:\windows 5. Type the following commands, pressing ENTER after each one. (Note that System.da0 and User.da0 contain the number zero.) attrib -h -r -s system.dat attrib -h -r -s system.da0 attrib -h -r -s user.dat attrib -h -r -s user.da0 6. Then type: copy system.da0 system.dat copy user.da0 user.dat 7. Restart your computer. Publication date: 10/16/2006 7-41 Varian 700-ES Series instruments 7.8 Contacting Technical Support If you confront problems that you cannot repair without assistance, you should send an email to: osi_tech.support@varianinc.com 7.7.1 Information required When requesting assistance you must include: 1. Instrument serial number 2. Instrument Model, include options. 3. Software product and version and build number (see ..Help...About) 4. EPROM versions (firmware versions) 5. Computer brand, type and speed, and unusual features. For instance, network adapters, sound cards, RAM, software applications. 6. Detailed problem description 7. Copies of related files for diagnosis 8. Accessories involved 9. Applications, methodology, types of samples. 10. Details of what has been done so far to fix the problem 11. General lab conditions. Include power supply outlets, is it a “clean room”, other instruments in the room. 7.7.2 Related files When sending files to tech support for review you should include: 1. A copy of the affected worksheet 2. A copy of the error.log 3. A copy of the wavelengthcalibration.txt file 4. For the Varian 710-ES Series instrument A copy of the AttenuatorCalibration.txt and ShutterCalibration.txt files 5. A copy of the stored dark current file 6. You should include the debug log only if asked by osi_tech.support@varianinc.com 7-42 Publication date: 10/16/2006