PRACTICAL MANUAL Molecular Biology Course Instructor Dr. Pakeeza A. Shaiq Teacher’s Assistant Ms. Sadaf Naz Mr. Muhammad Mazhar University Institute of Biochemistry and Biotechnology Practical#1 Laboratory Ethics General Rules and Regulations: It is necessary for everyone working in the lab to observe safety rules and regulations of each laboratory. Below are general safety guidelines applicable to most laboratories: 1. Personal Protective Equipment (PPE): Lab coats or disposable aprons should be worn in the lab to protect you and your clothing from contamination. Lab coats should not be worn outside the laboratory. Lab footwear should consist of normal closed shoes to protect all areas of the foot from possible puncture from sharp objects or broken glass and from contamination from corrosive reagents or infectious materials. Gloves should be worn for handling chemicals, blood and specimens. Protective eyewear such as goggles and masks may need to be worn when contact with hazardous aerosols, caustic chemicals or reagents. 2. Eating, drinking, smoking, applying cosmetics and handling contact lenses are prohibited in areas where specimens are handled. 3. Smoking is prohibited in all laboratory areas. 4. Food and drink should not be stored in refrigerators, freezers, cabinets or on shelves where chemicals, blood or other potentially infectious materials are stored. 5. Long hair, ties, scarves and earrings should be secured. 6. Never mouth pipette, mechanical pipetting devices must be used for pipetting all liquids. 7. Frequent hand washing is an important safety precaution, which should be practiced before and after every experiment. 8. Label all storage areas appropriately and keep all chemicals in properly labelled containers. 9. Note expiry dates on chemicals. 10. Always return reagents to their proper place in the laboratory immediately after use. 11. Always return the cap to the proper bottle and ensure that it is secured on the bottle. 12. Never return unused chemicals to a reagent bottle. Doing so might result in the contamination of the reagent. 13. When heating or carrying out reactions in a test tube, never point the mouth of the tube at your neighbor or yourself. 14. Never taste a chemical; never smell a chemical unless instructed to do so. If instructed to smell a chemical, fan vapors toward your nose and inhale cautiously. 15. Work in a fume hood when handling toxic and fume producing chemicals. 16. Handle and store glassware carefully so as not to damage it or yourself. 17. When inserting glass tubing into rubber stoppers, corks or when placing rubber tubing on glass hose connections protect hands with a heavy glove and lubricate tubing or stopper. 18. When dealing with broken glass wear hand protection when picking up the pieces. Use a broom to sweep small pieces into a dustpan and store glass pieces in a designated bin for broken glass. 19. Paper, broken glass, stoppers, rubber tubing, etc., are to be kept out of sinks at all times to minimize the possibility of clogged drains. Such items are to be kept away from areas where they might fall into sinks or drains. 20. Do not work with flammable solvents/chemicals near open flames (e.g. a gasburner) at the same time in the same laboratory. 21. Warning signs should be posted to alert attentions in the work area when unusual hazards, such as radiation, laser operations, dangerous chemicals, biological hazards, or other special hazards exist. 22. Before leaving the laboratory electrical equipments should be turned off and gas burners extinguished. No tap water should be left running. Hazard Symbols Biosafety Cabinets Introduction: Biosafety cabinets also known as biological safety cabinets (BSCs) are an enclosed, ventilated hood or workspace that allow for the safe handling of pathogens, contaminants or other potentially hazardous materials. The primary purpose of a biosafety cabinet is to protect the operator and the surrounding environment from biological contaminants and other hazardous materials. These cabinets are designed to provide various levels of protection. There are various classes of biological safety cabinets, each defined by the required level of biosafety and containment. Classes of biosafety cabinets: Class Ι: The Class I biosafety cabinet provides personnel and environmental protection, but no product protection. It has a HEPA filter in the exhaust system to provide containment and environmental protection. This older class of biosafety cabinet is rarely used. Class ΙΙ: The Class II biosafety cabinet provides protection to the user, the experimental material and the environment. Air flow is drawn from the room around the operator into the front grille of the cabinet, which provides personnel protection. In addition, the downward laminar flow of HEPA-filtered air provides protection for experimental material inside the cabinet. Because cabinet air has passed through the exhaust HEPA filter, it is contaminant-free, providing environmental protection, and may be recirculated back into the laboratory. (Class II Type A) or ducted out of the building (Class II Type B). Class ΙΙΙ: The Class III biological safety cabinet is most suitable for work with bio-hazardous agents requiring high contain (biosafety level 3 or 4). The Class III cabinet is completely enclosed, HEPA filter-ventilated cabinet fitted with glove ports and decontamination capabilities for entry and exit of material. It offers the highest degree of personnel and environmental protection from infectious aerosols. Clean bench & fume hood: Clean Benches and Fume Hoods are Not Biological Safety Cabinets. Fume hoods are ventilated enclosures that remove hazardous chemical fumes and volatile vapors from the laboratory, providing personnel protection only, hence should not be used for biohazard substances. Whereas, a clean bench provides a space to work with a product or specimen where it will be protected from contamination by particulates such as microorganisms. This is accomplished by the laminar flow of clean air from a HEPA filter, which is blown across the workspace and out toward the user and the lab. Thus, clean benches should not be used when working with potentially infectious materials, chemical hazards or radioactivity. Biosafety Levels Definition: A biosafety level (BSL) is a set of bio-containment precautions required to isolate dangerous biological agents in an enclosed laboratory facility. Biosafety levels: Biosafety Levels 1-4 are combinations of laboratory practices and techniques, safety equipment and facilities. All of these levels are appropriate for the biohazard posed by the agents used in researches and experiments. Biosafety level 1: Biosafety level one, the lowest level, applies to work with agents that usually pose a minimal potential threat to laboratory workers and the environment and do not consistently cause disease in healthy adults. An example of a microbe that is typically worked with at a BSL-1 is a nonpathogenic strain of E. coli. Precautions of biosafety level 1: 1. Normal laboratory personal protective equipment is generally worn, consisting of eye protection, gloves and a lab coat or gown. 2. Standard microbiological practices also require attention to personal hygiene, i.e., hand washing and a prohibition on eating, drinking or smoking in the lab. 3. Research with these agents is generally performed on standard open laboratory benches without the use of special containment equipment. 4. Labs are not usually isolated from the general building. 5. Standard microbiology practices are usually enough for protection such as mechanical pipetting only (no mouth pipetting allowed), safe sharps handling, avoidance of splashes or aerosols, and decontamination of all work surfaces when work is complete. 6. Decontamination of spills is done immediately, and all potentially infectious materials are decontaminated prior to disposal, generally by autoclaving. Biosafety level 2: Biosafety level two deals with agents associated with human disease, in other words, pathogenic or infectious organisms posing a moderate hazard. Example: Dealing with HIV when performing routine diagnostic procedures or work with clinical specimens. Precautions of biosafety level 2: 1. Appropriate personal protective equipment (PPE) must be worn, including lab coats and gloves. Eye protection and face shields can also be worn, as needed. 2. Laboratory personnel have specific training in handling pathogenic agents. 3. Access to the laboratory is limited when work is being conducted and the lab has self-closing, lockable doors. 4. Extreme precautions are taken with contaminated sharp items. 5. Class II biological safety cabinet is highly recommended for work involving these agents. Biosafety level 3: BSL-3 is suitable for work with infectious agents which may cause serious or potentially lethal diseases as a result of exposure by the inhalation route. Example: M. tuberculosis. Precautions of biosafety level 3: 1. Solid-front wraparound gowns, scrub suits or coveralls are often required. 2. BSL-3 laboratories are located in a unique high containment building that also houses the BSL-4 laboratory that have a double-door entry. 3. Exhaust air is not recirculated to other rooms. 4. Standard microbiological practices are the same as for BSL-1 and BSL-2 laboratories. 5. Class ΙΙ biological safety cabinets are suitable in BSL-3 laboratories. 6. Additional personnel protective devices may be worn, such as respirators. Biosafety level 4: BSL 4 deals with extremely dangerous agents and pose a high risk of life-threatening disease. Examples are the Ebola virus, the Lassa virus, and any agent with unknown risks of pathogenicity and transmission. Precautions of biosafety level 4: 1. To the BSL 3 practices, we add requirements for complete clothing change before entry, a shower on exit and decontamination of all materials prior to leaving the facility. 2. Personnel must wear appropriate personal protective equipment from prior BSL levels, as well as a full body, air-supplied, positive pressure suit. 3. The BSL 4 laboratory contains a class III biological safety cabinet. BSL 4 laboratories are in separate buildings or a totally isolated zone with dedicated supply and exhaust ventilation. Practical No 2 Isolation of genomic DNA from plants by CTAB method Principle: extraction of genomic DNA from plant material requires cell lysis, inactivation of nucleases and separation of desired genomic DNA from cellular debris. The cetyl trimethyl ammonium bromide (ctab) protocol developed by muray and thompson (1980) is appropriate for extraction of genomic DNA from plants and from plant derived food stuff and is particularly useful for the elimination of polysaccharides and polyphenolic compounds otherwise affecting the DNA purity and quality. Plant cells can be lysed with the ionic detergent ctab, which forms an insoluble complex with nucleic acids in a low-salt environment. Under these conditions, polysaccharides, phenolic compounds and other contaminants remain in the supernatant and can be washed away. The DNA complex is solubilised by raising the salt concentration and precipitated with ethanol or isopropanol. Extraction of genomic DNA from plants involves three steps Lysis of cell membrane Extraction Precipitation 1. Lysis of cell membrane: the first step of the DNA extraction is the rupture of the cell and nucleus wall. For this purpose, the homogenized sample is first treated with the extraction buffer containing EDTA, tris/hcl and ctab. The lysis of the membranes is accomplished by the detergent (ctab) contained in the extraction buffer. When the cell membrane is exposed to the ctab extraction buffer, the detergent captures the lipids and the proteins allowing the release of the genomic DNA. In a specific salt (nacl) concentration, the detergent forms an insoluble complex with the nucleic acids. EDTA is a chelating component that among other metals bind magnesium. Magnesium is a cofactor for DNAse. By binding mg with EDTA, the activity of present DNAse is decreased. Tris/hcl gives the solution a ph buffering capacity (a low or high ph damages DNA). After the cell and organelle membranes (such as those around the mitochondria and chloroplasts) have been broken apart, the purification of DNA is performed. Extraction: in this step, polysaccharides, phenolic compounds, proteins and other cell lysates dissolved in the aqueous solution are separated from the ctab nucleic acid complex. Under low salt concentration, the contaminants of the nucleic acid complex do not precipitate and can be removed by extraction of the aqueous solution with chloroform. The chloroform denatures the proteins and facilitates the separation of the aqueous and organic phases. Once the nucleic acid complex has been purified, precipitation can be accomplished. Precipitation: in this final stage, the nucleic acid is liberated from the detergent. For this purpose, the aqueous solution is first treated with a precipitation solution comprising of sodium acetate, which precipitates the nucleic acid. Under these conditions, the detergent, which is more stable in alcohol than in water, can be washed out, while the nucleic acid precipitates. The successive treatment with 70% ethanol allows an additional purification, or wash, of the nucleic acid from the remaining salt Protocol for genomic DNA isolation using ctab Take 2-3 leaves of plant tissue, homogenate it in sterilized pistil and motar with 2ml ctab buffer (preheated at 65°c for 15 min). Ctab buffer contain (100mm tris hcl ph 8.0, 2% (w/v) ctab, 20mm EDTA, 1.4m nacl, 1% ctab and 1% β-mercaptoethanol). Transfer 200 mg of homogenate to sterile eppendrof tube, and add 600 µl CTAB and mix vigorously and incubate it in water bath at 65°c for 1hr Bring down the sample at room temperature, add 600µl of chloroform isoamyl alcohol mixture (24:1), mix gently by inverting and spin at 10,000 rpm for 10 minutes Transfer the supernatant to a new eppendrof tube and add 600µl chroform isoamyl alcohol mixture, mix and spin at 10,000 rpm for 10 minutes Collect the supernatant to new eppendrof tube and add 600µl of cold isopropanol. Mix gently to precipitate the nucleic acid. Spin for 5-10minutes at 10,000 rpm. Wash the pellet with 700µl of 70% ethanol. Decant and dry the pellet at room temperature. Dissolve in 50µl of 0.1xte+rnase. Incubate for 1 hr at 37 °c. DNA check: Load 2-4 µl of isolated plant genomic DNA in 1% agrose gel and determine the quality and yield. Agarose gel electrophoresis of genomic DNA isolated from plant Practical No 3 Extraction of genomic DNA from Animal tissue DNA extraction from blood The white blood cells (wbc) of peripheral blood are usually the most convenient source of human genomic DNA for DNA analysis with respect to haemoglobinopathies. It is estimated that 10 ml of whole blood yield approximately 250 μg of DNA, more than sufficient for complete analysis of globin genes with the methods that are currently available (ie based on pcr). DNA is an extremely stable molecule, but enzymes which catalyze the breakdown of nucleic acids (nucleases) are found in all cells. In intact cells the DNA is found in the nucleus and thus is protected from the action of nucleases which are abundant in the lysosomes in the cytoplasm. However when cells are lysed, the membranes of the cell compartments are disrupted, allowing nucleases to come in to contact with the DNA. Thus DNA extraction uses buffers which contain inhibitors of nuclease activity i.e.: (EDTA) Preparation of buffers Cell lysis buffer (clb) The final volume will be 1000ml that contains 100mm tris-hcl (ph 7.6), 5mm mgcl2, 7mm kcl, 300mm sucrose and 1% triton x-100. Nucleus lysis buffer (nlb) The nucleus lysis buffer contains 100mm tris-hcl, 12mm sodium acetate/ potassium acetate or 10mm sodium citrate, 2% sodium dodecyl sulphate (sds) and 10mm EDTA. DNA suspending buffer (te buffer) 10mm tris-hcl and 1mm EDTA. For 1 litter 5m sodium chloride (nacl) solution Dissolve 29.2 gm of nacl in 100ml distill water. All the buffer must be sterilized prior use and before starting DNA extraction, chloroform and absolute ethanol (100%) should be kept on -20oc before starting DNA extraction. Eukaryotic DNA extraction protocol: Take 600 μl of clb (cell lysis buffer) in pre-labeled eppendorf tube. o Add 300 μl of EDTA/heparin blood and mix gently by inverting 4-6 times. o Centrifuge at 4000 rpm for 5 min. o Decant the supernatant and wash the pallet again with clb 400 μl. o Centrifuge at 5000 rpm for 5 min. o Repeat washing (3-4 times) until all the hemoglobin is washed out. o Add 450 μl nlb (nuclear lysis buffer) and 100 μl of saturated nacl to dried pallet. o Add 550 μl pre chilled chloroform. Centrifuge at 5000 rpm for 3 min. Two separate layers will be formed. o Withdraw the supernatant without touching the central layer to new labeled eppendorf tube. This supernatant will be processed for further DNA extraction. o Add 1000 μl pre-chilled ethanol. o Place tubes at -20°c for 30 min and then centrifuge at 14000rpm for 6 min and discard ethanol slowly. o Add 1000 μl of 70% ethanol. o Centrifuge at 14000rpm for 6 min. Decant ethanol and let tube dry completely. o Add 100 μl of t.e buffer and store at 37°c for 24hrs. o After that check DNA on 0.8% agarose gel and quantify using nanodrop. o Store DNA at-20°c DNA extraction buffers Cell lysis buffer: Tris hcl 100ml Mgcl2 1.01g Kcl 0.521g Sucrose 109.54g Triton x100 10ml Dist. H2o make volume up to 1000 Nuclear lysis buffer: Tris hcl 100ml Sodium citrate 2.9g EDTA 20ml Sds 20g Dist. H2o make volume up to 1000 T.e buffer: Tris hcl 10ml EDTA 2ml Dist. H2o make volume up to 1000 Nacl solution: 350 g in 1000ml Tris hcl: 1. Tris 121.1g 2. Dist. H2o make volume up to 1000 3. Hcl make ph 7.6 EDTA solution: 1. EDTA 186.1g Dist. H2o make volume up to 1000 Naoh make ph 8 What each reagent is for? Lysis buffer: A lysis buffer is a buffer solution used for the purpose of breaking open cells for use in molecular biology experiments CLB Breaking the cell membranes open to expose the DNA along with the cytoplasm within the cell NLB Breaking the nucleus Triton x-100 and sds : Detergents Are added to break up membrane structures. EDTA: EDTA is responsible for chelation of divalent ions EDTA helps to stop DNAses (DNA cutting enzymes) present in the cytoplasm from acting on the exposed DNA Mg2+ is an important factor for activity of DNAses. EDTA deprives the enzyme of this co-factor and renders it inactive DNAses are enzymes that “chew up” DNA and thus reduces the yield of genomic DNA. So it’s important to keep them from acting on DNA of interest. Tris-hcl: Maintains ph Salt solution: The solution is treated with salt solution to make debris such as broken proteins, lipids and rna to clump together. Sodium chloride helps to remove proteins that are bound to the DNA. Chloroform: Used to help separate proteins, lipids and polysaccharides from nucleic acids in the cell. Chloroform and water separates into two distinct phases. Lower phase will be chloroform containing proteins,lipids and polysaccharides Chilled ethanol : DNA doesn’t dissolve in ethanol. It will aggregate together, giving a pellet upon centrifugation Used to separate pure DNA Colder temperature reduces activity of enzymes that can break down DNA. 70% ethanol: This step is to wash any residual salt away from the pelleted DNA. T.E buffer : Used to solubilize DNA while protecting it from degradation Practical 4 Spectrophotometric quantification of nucleic acids After isolation of DNA, quantification and analysis of quality are necessary to ascertain the approximate quantity of DNA obtained and the suitability of DNA sample for further analysis. This is important for many applications including digestion of DNA by restriction enzymes or PCR amplification of target DNA. The most commonly used methodologies for quantifying the amount of nucleic acid in a preparation are: Gel electrophoresis. Spectrophotometric analysis. Spectrophotometric analysis: Spectrophotometer: Instrument having two important parts: 1. Spectrometer- emit light of specific wavelength UV (10- 400 nm) Visible (400-780nm) 2. Photometer- measures the intensity of light Purines and pyrmidines in nucleic acid show absorption maxima around 260nm (eg.,dATP: 259nm; dCTP: 272nm; dTTP: 247nm). Spectrophotometers are commonly used to determine the concentration of DNA. Inside a spectrophotometer, a sample is exposed to ultraviolet light at 260 nm, and a photo-detector measures the light that passes through the sample. The more light absorbed by the sample, the higher the nucleic acid concentration in the sample. Beer-lambert law: According to this absorbance is directly proportional to path length and concentration of solution. A = εlc Where: A=absorbance in terms of optical density (od) Ε=extinction coefficient L=path length C=concentration Nucleic acid type Value of ε DsDNA 50g/ml SsDNA or RNA 40 g/ml Ss oligonucleotides 33 g/ml Concentration of double stranded DNA in g/ml = absorbance at 260nm x 50 x dilution factor. Concentration of single-stranded DNA and RNA in g/ml = absorbance at 260nm x 40 x dilution factor. Sample purity: It is common for nucleic acid samples to be contaminated with other molecules (i.e. Proteins, organic compounds, other). The ratio of the absorbance at 260 and 280nm (a260/280) is used to assess the purity o nucleic acids. A ratio between 1.8 and 2.0 denotes that the absorption in the UV range is due to nucleic acids. A ratio lower than 1.8 indicates the presence of proteins and/or other UV absorbers. A ratio higher than 2.0 indicates that the samples may be contaminated with chloroform or phenol. In either case (<1.8 or >2.0) it is advisable to re-precipitate the DNA. Demerits of spectrophotometric determination: This method is however limited by the quantity of DNA and the purity of the preparation. Accurat analysis of the DNA preparation may be impeded by the presence of impurities in the sample or if the amount o DNA is too little. In the estimation of total genomic DNA, for example, the presence of RNA, sheared DNA etc Could interfere with the accurate estimation of total high molecular weight genomic DNA. Protocol: 1. Turn the spectrometer on 2. Turn on the UV lamp 20 min before you will take your readings, the visible light lamps can be used immediately, but UV lamp takes a while to become steady. The amount of warm up time needed depends on the lamp and the spectrometer 3. Your sample will be DNA or RNA in water or buffer. 4. Put the sample and the blank in a matched set of quartz cuvettes 5. Set the wavelength to 260 nm 6. Blank the machine 7. Read o.d. of sample at 260 nm 8. Set wavelength to 280nm. Reblank and read the o.d. at 280 9. Calculate the concentration of DNA using following information a. 1a260 unit of double stranded DNA= 50µg (50 µg/ml has an o.d. of 1 at 260 nm) 10. DNA concentration (µg/ml) = od260 × (dilution factor) ×50 µg DNA/ml 1od260 unit. 11. Calculate the yield of your preparation by using formula Yield= (DNA concentration in µg/ml x total volume of solution in ml) 12. Estimate the purity of preparation by figuring the 260/280 ratios. The ratio b/w the readings at 260 nm and 280nm gives an estimate of the purity of the nucleic acid. Pure preparations of DNA should have a 260/280 ratio of 1.8, RNA a ratio of 2. Practical 5 Agarose Gel Electrophoresis Principle: Agarose gel electrophoresis separates DNA fragments according to their size. An electric current is used to move the DNA molecules across an agarose gel, which is a polysaccharide matrix that functions as a sieve to help "catch" the molecules as they are transported by the electric current. The phosphate molecules that make up the backbone of DNA molecules have a high negative charge. When DNA is placed on a field with an electric current, these negatively charged DNA molecules migrate toward the positive end of the field, which in this case is an agarose gel immersed in a buffer bath. The agarose gel is a cross-linked matrix i.e., a three-dimensional mesh or screen. The DNA molecules are pulled to the positive end by the current, but they encounter resistance from this agarose mesh. The smaller molecules are able to navigate the mesh faster than the larger ones. This is how agarose electrophoresis separates different DNA molecules according to their size. The gel is stained with ethidium bromide so as to visualize these DNA molecules resolved into bands along the gel. Ethidium bromide is an intercalating dye, which intercalate between the bases that are stacked in the center of the DNA helix. One ethidium bromide molecule binds to one base. As each dye molecule binds to the bases the helix is unwound to accommodate the stain from the dye. Closed circular DNA is constrained and cannot withstand as much twisting strain as can linear DNA, so circular DNA cannot bind as much dye as can linear DNA. Unknown DNA samples are typically run on the same gel with a "ladder." a ladder is a sample of DNA where the sizes of the bands are known. Unknown fragments are compared with the ladder fragments (size known) to determine the approximate size of the unknown DNA bands. Approximately 10ng is visible in a single band on a horizontal agarose gel. Materials: •Agarose •Tae Buffer •Gel Casting Tray, Comb, Power Pack •Sample DNA •Loading Dye •Sterile Micro Tips •ETBR Staining Solution •UV Transilluminator Or Gel Documentation System Instructions: For casting gel, agarose powder is mixed with electrophoresis buffer (TAE) to the desired concentration, and then heated in a microwave oven until completely melted. After cooling the solution to about 60°C, it is poured into a casting tray containing a comb and allowed to solidify at room temperature for nearly 45 min. After the gel has solidified, the comb is removed, using care not to rip the bottom of the wells. The gel, still in its plastic tray, is inserted horizontally into the electrophoresis chamber and just immersed with buffer (TAE). DNA samples mixed with loading buffer are then pipetted into the sample wells, the lid and power leads are placed on the apparatus, and a current is applied. The current flow is confirmed by observing bubbles coming off the electrodes. DNA will migrate towards the positive electrode, which is usually colored red. The distance DNA has migrated in the gel can be judged by visually monitoring migration of the tracking dyes. Bromophenol blue and xylene cyanole dyes migrate through agarose gels at roughly the same rate as double-stranded DNA fragments of 300 and 4000 bp, respectively. Preparation of 1% agarose gel: Weigh 1g agarose; add in 100 ml 1xTAE and melt agarose in a microwave oven for 2-3 min. Cool down to about 45 to 50°C (bearable warmth) and pour into the gel platform with the comb in position. Running Gel: After solidification of the gel (approx. 45 min), place the gel in a gel tank with 1 x TAE buffer. Buffer should be filled to the surface of the gel. Load the samples in the well and run the gel at 80V till the blue dye runs to the end. Staining the gel: Prepare staining solution by adding 10 µl of 10 mg/ml stock of ethidium bromide in 100 ml of distilled water. Place the gel in staining solution for 30 min and view the gel in UV transilluminator. Fig 1: agarose gel electrophoresis method Practical 6 UV Visualization and Analysis of Gel Results If the gel is done correctly, upon visualization there will be a row of well-defined bands in the gel. However, some errors can cause smearing, and the bands will not be distinguishable. Gel improperly prepared Smearing can be caused by an improperly prepared gel. If the gel is not poured correctly, it will not polymerize or solidify evenly, thus causing the molecules to smear. Overloading the wells The sample of the molecule is placed into wells at one end of the gel. If these wells are filled too much, or if the sample is not properly diluted, the excess sample may smear across the gel. In addition, if the gel is moved after the sample is placed in the well, it can cause the sample to spill out of the well. This can also cause smearing. Contamination According to "nucleic acid gel electrophoresis," another cause of smearing is contamination of the sample. For example, a DNA sample may be contaminated with a protein, or a protein sample may be contaminated by lipids or fats. Determining the DNA fragment length A 'reference ladder' can also be run in the gel. This contains a mixture of DNA fragments of known size. Comparing the bands in your DNA sample with the bands in the reference ladder allows you to work out how big the DNA fragments are in a particular band. DNA size is measured in base pairs (bp), or kilo-base pairs. Gel documentation systems Gel documentation, or gel imaging, systems are used to record and measure labeled nucleic acid and protein in various types of media such as agarose, acrylamide or cellulose. Systems come in a variety of configurations depending on throughput and sample type. Standard operating procedure (sop) for using gel documentation system (gel doc system) and handling of agarose gels containing ethidium bromide (etbr) 1. Do not use gloves to open and close the doors of common instrument room as well as any other rooms in bsbe. 2. Designate a separate plastic tray for carrying gel containing etbr, gloves, tissue paper and 70% ethanol / distilled water. 3. Before using gel doc system, clean the surface of trans-illuminator with 70 % ethanol or distilled water. 4. Wear gloves to handle the gel containing etbr and then place the gel on to the surface of trans-illuminator. 5. Remove the gloves and close the door of gel doc system. 6. Document the gel picture on the computer without wearing the gloves. 7. At any moment of time, computer, keyboard, mouse and gel doc system should not be used with gloves. 8. Wear gloves to remove the gel containing etbr and clean the surface of trans-illuminator. 9. Remove the gloves and close the door of gel doc system. 10. Make an appropriate entry in the log book of gel doc system. 11. Carry all the material in the designated plastic tray. 12. Report immediately to concerned in-charge for problem regarding system, log book etc. Practical 7 Melting kinetics of DNA Melting temperature of DNA: (melting temperature) temperature at which half of the DNA is melted Melting temperature of DNA is affected by 3 main factors Nucleotide content of DNA molecule Length of DNA molecule Ionic strength of the DNA molecule Refers to the separation of 2 stands Breaking of h-bonds + base stacking interactions Also referred to as denaturation Thermal energy > intermolecular bond energy (denaturation) A:T separates more rapidly than G:C Tm=t at which half of the DNA is melted How to calculate melting temperature of DNA Tm = 4(G+C) +2(A+T) Tm =4(6) +2(5) (suppose ATCGATATAACGGC) Tm=24+10=34 If the GC content of a DNA molecule is 60%, what are the percentages of the four bases? Re-naturation of DNA Bringing the two strands together by simply lowering the temperature Temperature can be lower to anneal DNA strand Reform h-bond and base stacking Rapid cooling result in mismatched base pairing Thermal energy now too low for DNA to melt and find their proper complements Practical No 8 Protein Estimation by Lowry method The Lowry assay (1951) is an often-cited general use protein assay. For some time it was the method of choice for accurate protein determination for cell fractions, chromatography fractions, enzyme preparations, and so on. The bicinchoninic acid (BCA) assay is based on the same princple and can be done in one step, therefore it has been suggested (Stoscheck, 1990) that the 2-step Lowry method is outdated. Principle The principle behind the Lowry method of determining protein concentrations lies in the reactivity of the peptide nitrogen[s] with the copper [II] ions under alkaline conditions. Under alkaline conditions the divalent copper ion forms a complex with peptide bonds in which it is reduced to a monovalent ion. Monovalent copper ion and the radical groups of tyrosine, tryptophan, and cysteine react with Folin reagent to produce an unstable product that becomes reduced to molybdenum/tungsten blue. Equipment In addition to standard liquid handling supplies a spectrophotometer with infrared lamp and filter is required. Glass or polystyrene (cheap) cuvettes may be used. Procedure: Reagents A. 2% Na2CO3 in 1% NaOH B. 0.5% CuSO4 in 1 % Na-K Tartrate Reagent-I :50ml A + 1ml B Reagent-II :1 part Folin-Phenol [2 N]: 1 part water BSA Standard - 1 mg/ ml BSA is the universally accepted reference protein for total protein quantitation Protocol: 0.2 ml of BSA working standard in 5 test tubes and make up to 1ml using distilled water. The test tube with 1 ml distilled water serve as blank. Add 4.5 ml of Reagent I and incubate for 10 minutes. After incubation add 0.5 ml of reagent II and incubate for 30 minutes . Estimate the amount of protein present in the given sample from the standard graph. Measure the absorbance between 650-750 nm and plot the standard graph . Estimate the amount of protein present in the given sample from the standard graph. Analysis Prepare a standard curve of absorbance versus micrograms protein (or vice versa), and determine amounts from the curve. Determine concentrations of original samples from the amount protein, volume/sample, and dilution factor, if any. Practical 9 Bradford protein assay The Bradford assay is very fast and uses about the same amount of protein as the Lowry assay. It is fairly accurate and samples that are out of range can be retested within minutes. The Bradford is recommended for general use, especially for determining protein content of cell fractions and assesing protein concentrations for gel electrophoresis.` Principle: The Bradford Protein Assay measures the concentration protein by adding Coomassie dye to the sample under acidic conditions. When proteins bind with the Coomassie dye, the sample changes color from brown to blue. The level of blue can then be measured using a spectrophotometer to determine the concentration of protein in the sample. The assay is based on the observation that the absorbance maximum for an acidic solution of Coomassie Brilliant Blue G-250 shifts from 465 nm to 595 nm when binding to protein occurs. Both hydrophobic and ionic interactions stabilize the anionic form of the dye, causing a visible color change. The assay is useful since the extinction coefficient of a dye-albumin complex solution is constant over a 10fold concentration range. Equipment In addition to standard liquid handling supplies a visible light spectrophotometer is needed, with maximum transmission in the region of 595 nm, on the border of the visible spectrum (no special lamp or filter usually needed). Glass or polystyrene (cheap) cuvettes may be used, however the color reagent stains both. Disposable cuvettes are recommended. Procedure Reagents Bradford reagent: Dissolve 100 mg Coomassie Brilliant Blue G-250 in 50 ml 95% ethanol, add 100 ml 85% (w/v) phosphoric acid. Dilute to 1 liter when the dye has completely dissolved, and filter through Whatman #1 paper just before use. (Optional) 1 M NaOH (to be used if samples are not readily soluble in the color reagent). Protocol Warm up the spectrophotometer before use. Dilute unknowns if necessary to obtain between 5 and 100 µg protein in at least one assay tube containing 100 µl sample If desirred, add an equal volume of 1 M NaOH to each sample and vortex.The addition of 1 M NaOH was suggested by Stoscheck (1990) to allow the solubilization of membrane proteins and reduce the protein-toprotein variation in color yield.Add NaOH to standards as well if this option is used. Prepare standards containing a range of 5 to 100 micrograms protein (albumin or gamma globulin are recommended) in 100 µl volume. See how to set up an assay for suggestions as to setting up the standards. Add 5 ml dye reagent and incubate 5 min. Measure the absorbance at 595 nm. Analysis Prepare a standard curve of absorbance versus micrograms protein and determine amounts from the curve. Determine concentrations of original samples from the amount protein, volume/sample, and dilution factor, if any Practical 10 Protein separation by SDS-PAGE Introduction: The smaller molecules migrate faster due to less resistance during electrophoresis. The structure and the charge of the proteins also influence the rate of migration. Sodium dodecyl sulphate and polyacrylamide eliminate the influence of structure and charge of the proteins, and the proteins are separated based on the length of the polypeptide chain Principle of SDS-PAGE The principle of SDS-PAGE states that a charged molecule migrates to the electrode with the opposite sign when placed in an electric field. The separation of the charged molecules depends upon the relative mobility of charged species. Role of SDS in SDS-PAGE SDS is a detergent present in the SDS-PAGE sample buffer. SDS along with some reducing agents function to break the disulphide bonds of proteins disrupting the tertiary structure of proteins. Materials Required Power Supplies: It is used to convert the AC current to DC current. Gels: These are either prepared in the laboratory or precast gels are purchased from the market. Electrophoresis Chambers: The chambers that can fit the SDS-PAGE gels should be used. Protein Samples: The protein is diluted using SDS-PAGE sample buffer and boiled for 10 minutes. A reducing agent such as dithiothreitol or 2-mercaptoethanol is also added to reduce the disulfide linkages to prevent any tertiary protein folding. Running Buffer: The protein samples loaded on the gel are run in SDS-PAGE running buffer. Staining and Destaining Buffer: The gel is stained with Coomassie Stain Solution. The gel is then destained with the destaining solution. Protein bands are then visible under naked eyes. Protein Ladder: A reference protein ladder is used to determine the location of the protein of interest, based on the molecular size. Protocol of SDS-PAGE Preparation of the Gel All the reagents are combined, except TEMED, for the preparation of gel. When the gel is ready to be poured, add TEMED. The separating gel is poured in the casting chamber. Add butanol before polymerization to remove the unwanted air bubbles present. The comb is inserted in the spaces between the glass plate. The polymerized gel is known as the “gel cassette” Sample Preparation Boil some water in a beaker. Add 2-mercaptoethanol to the sample buffer. Place the buffer solution in micro-centrifuge tubes and add protein sample to it. Take MW markers in separate tubes. Boil the samples for less than 5 minutes to completely denature the proteins. Electrophoresis The gel cassette is removed from the casting stand and placed in the electrode assembly. The electrode assembly is fixed in the clamp stand. 1x electrophoresis buffer is poured in the opening of the casting frame to fill the wells of the gel. Pipette 30ml of the denatured sample in the well. The tank is then covered with a lid and the unit is connected to a power supply. The sample is allowed to run at 30mA for about 1 hour. The bands are then seen under UV light. Applications of SDS-PAGE The applications of SDS-PAGE are as follows: 1. It is used to measure the molecular weight of the molecules. 2. It is used to compare the polypeptide composition of different structures. 3. It is used to estimate the purity of the proteins. 4. It is used in Western Blotting and protein ubi-quitination. 5. Analyzing the size and number of polypeptide subunits. 6. To analyze post-translational modifications. Practical 11 Centrifugation Centrifugation is a technique used for the separation of particles from a solution according to their size, shape, density, viscosity of the medium and rotor speed. The particles are suspended in a liquid medium and placed in a centrifuge tube. The tube is then placed in a rotor and spun at a define speed. Separation through sedimentation could be done naturally with the earth gravity, nevertheless, it would take ages. Centrifugation is making that natural process much faster. A centrifuge is a device that separates particles from a solution through use of a rotor. Principle As a rotor spins in a centrifuge, a centrifugal force is applied to each particle in the sample; the particle will then sediment at the rate that is proportional to the centrifugal force applied to it. The viscosity of the sample solution and the physical properties of the particles also affect the sedimentation rate of each particle. At a fixed centrifugal force and liquid viscosity, the sedimentation rate of a particle is proportional to its size (molecular weight) and to the difference between the particle density and the density of the solution. In a solution, particles whose density is higher than that of the solvent sink (sediment), and particles that are lighter than it float to the top. The greater the difference in density, the faster they move. If there is no difference in density (isopycnic conditions), the particles stay steady. Types of centrifuge Low speed centrifuge 1) Most laboratories have a standard low-speed centrifuge used for routine sedimentation of heavy particles 2) The low-speed centrifuge has a maximum speed of 4000-5000rpm 3) These instruments usually operate at room temperatures with no means of temperature control. 4) Two types of rotors are used in it, Fixed angle Swinging bucket. 5) It is used for sedimentation of red blood cells until the particles are tightly packed into a pellet and supernatant is separated by decantation. High speed centrifuge 1. High-speed centrifuges are used in more sophisticated biochemical applications, higher speeds and temperature control of the rotor chamber are essential. 2. The high-speed centrifuge has a maximum speed of 15,000 – 20,000 RPM 3. The operator of this instrument can carefully control speed and temperature which is required for sensitive biological samples. 4. Three types of rotors are available for high-speed centrifugation Fixed angle Swinging bucket Vertical rotors Ultracentrifuge 1. It is the most sophisticated instrument. 2. Ultracentrifuge has a maximum speed of 65,000 RPM (100,000’s x g). 3. Intense heat is generated due to high speed thus the spinning chambers must be refrigerated and kept at a high vacuum. 4. It is used for both preparative work and analytical work. Types of centrifugation Differential Centrifugation The simplest form of separation by centrifugation is differential centrifugation, sometimes called differential pelleting. Particles of different densities or sizes in a suspension will sediment at different rates, with the larger and denser particles sedimenting faster. These sedimentation rates can be increased by using centrifugal force. A suspension of cells subjected to a series of increasing centrifugal force cycles will yield a series of pellets containing cells of decreasing sedimentation rate. Practical 12 Introduction to PCR Thermocycler and Sanger Sequencing PCR is a laboratory method used for making a very large number of copies of short sections of DNA from a very small sample of genetic material. This process is called "amplifying" the DNA and it enables specific genes of interest to be detected or measured. DNA is made up of repeating sequences of four bases – adenine, thymine, guanine, and cytosine. These sequences form two strands that are bound together in a double helix structure by hydrogen bonds (like a spiral staircase). Each half of the helix is a complement of the other. In humans, it is the difference in the sequence of these bases on each strand of DNA that leads to the uniqueness of each person's genetic makeup. The arrangement of the bases in each gene is used to produce RNA, which in turn produces a protein. There are about 25,000 genes in a human genome, and expression of these genes leads to the production of a large number of proteins that make up our bodies. The DNA of other organisms such as bacteria and viruses is also composed of thousands of different genes that code for their proteins. How is the method performed? PCR is carried out in several steps or "cycles" in an instrument called a thermo cycler. This instrument increases and decreases the temperature of the specimen at defined intervals during the procedure. 1. The first step or cycle of PCR is to separate the strands of DNA into two single strands by increasing the temperature of the sample that contains the DNA of interest. This is called "Denaturing" the DNA. 2. Once the strands separate, the sample is cooled slightly and forward and reverse primers are added and allowed to bind to the single DNA strands. Primers are short sequences of bases made specifically to recognize and bind to the section of DNA to be amplified, which are the very specific sequence of bases that are part of the gene or genes of interest this is called Annealing. Primers are called "forward" and "reverse" in reference to the direction that the bases within the section of DNA are copied. 3. After the two primers attach to each strand of the DNA, a DNA enzyme (frequently Taq polymerase) then copies the DNA sequence on each half of the helix from the forward to the reverse primer, forming two double stranded sections of DNA, each with one original half and one new half known as EXTENSION. Taq polymerase is an enzyme found in a bacterium (Thermues aquaticus) that grows in very hot water, such as in geysers or hot springs. Polymerases copy DNA (or RNA) to make new strands. The Taq polymerase is especially helpful for laboratory testing because (unlike many other enzymes) it does not break down at very high temperatures needed to do PCR. 4. When heat is applied again, each of the two double strands separate to make four single strands and, when cooled, the primers and polymerase act to make four double strand sections. The four strands becomes eight in the next cycle, eight become sixteen, and soon. 5. Within 30 to 40 cycles, as many as a billion copies of the original DNA section can be produced and are then available to be used in numerous molecular diagnostic tests. This process has been automated so that a billion copies of the original DNA can be produced within a few hours Details of thermal profile for DNA amplification using PCR General recepie and requirements 1. 2. 3. 4. 5. 6. DNA template 50ng/ul Primers; Forward and reverse Taq Polymerase dNTPs Monovalent cations (KCL) Divalent Cations (MgCl2) 7. PCR Buffer 8. ddH2O/PCR Water/Nuclease free water Precautions 1. Avoid contamination 2. Tm of primers should be calculated before starting PCR 3. Conc. of template DNA should be kept in mind Sanger Sequencing What is sequencing? DNA sequencing is the process of determining the sequence ofnucleotide bases (As, Ts, Cs, and Gs) in a piece of DNA. Today, with the right equipment and materials, sequencing a short piece of DNA is relatively straightforward. Sequencing an entire genome (all of an organism’s DNA) remains a complex task. It requires breaking the DNA of the genome into many smaller pieces, sequencing the pieces, and assembling the sequences into a single long "consensus." However, thanks to new methods that have been developed over the past two decades, genome sequencing is now much faster and less expensive than it was during the Human Genome Project. Sanger sequencing: The chain termination method Regions of DNA up to about 900900900 base pairs in length are routinely sequenced using a method called Sanger sequencing or the chain termination method. Sanger sequencing was developed by the British biochemist Fred Sanger and his colleagues in 1977. It’s also known as the “chain termination method,” was developed by the English biochemist Frederick Sanger and his colleagues in 1977. This method is designed for determining the sequence of nucleotide bases in a piece of DNA (commonly less than 1,000 bp in length). Sanger sequencing with 99.99% base accuracy is considered the “gold standard” for validating DNA sequences, including those already sequenced through next-generation sequencing (NGS). Sanger sequencing was used in the Human Genome Project to determine the sequences of relatively small fragments of human DNA (900 bp or less). These fragments were used to assemble larger DNA fragments and, eventually, entire chromosomes. Principle A DNA primer is attached by hybridization to the template strand and deoxynucleosides triphosphates (dNTPPs) are sequentially addedto the primer strand by DNA polymerase. The primer is designed for the known sequences at 3’ end of the template strand. M13 sequences is generally attached to 3’ end and the primer of thisM13 is made. The reaction mixture also contains dideoxynucleoside triphosphate (ddNTPs) along with usual dNTPs. If during replication ddNTPs is incorporated instead of usual dNTPs in thegrowing DNA strand then the replication stops at that nucleotide. The ddNTPs are analogue of dNTPs. ddNTPs lacks hydroxyl group (-OH) at c3 of ribose sugar, so it cannotmake phosphodiester bond with nest nucleotide, thus terminates thenucleotide chain. Respective ddNTPs of dNTPs terminates chain at their respective site. For example ddATP terminates at A site. Similarly ddCTP, ddGTP and ddTTP terminates at C, G and T site respectively. Ingredients for Sanger sequencing Sanger sequencing involves making many copies of a target DNA region. Its ingredients are similar to those needed for DNA replication in an organism, or for polymerase chain reaction (PCR), which copies DNA in vitro. They include: A DNA polymerase enzyme A primer, which is a short piece of single-stranded DNA that binds to the template DNA and acts as a "starter" for the polymerase The four DNA nucleotides (dATP, dTTP, dCTP, dGTP) The template DNA to be sequenced However, a Sanger sequencing reaction also contains a unique ingredient: Dideoxy, or chain-terminating, versions of all four nucleotides (ddATP, ddTTP,ddCTP, ddGTP), each labeled with a different color of dye Dideoxy nucleotides are similar to regular, or deoxy, nucleotides, but with one key difference: they lack a hydroxyl group on the 3’ carbon of the sugar ring. In a regular nucleotide, the 3’ hydroxyl group acts as a “hook," allowing a new nucleotide to be added to an existing chain.Once a dideoxy nucleotide has been added to the chain, there is no hydroxyl available and no further nucleotides can be added. The chain ends with the dideoxy nucleotide, which is marked with a particular color of dye depending on the base (A, T, C or G) that it carries. Sanger Sequencing Steps The Sanger Sequencing method consist of six steps; The double-stranded DNA (dsDNA) is denatured into two single-stranded DNA (ssDNA). A primer that corresponds to one end of the sequence is attached. Four polymerase solutions with four types of dNTPs but only onetype of ddNTP are added. The DNA synthesis reaction initiates and the chain extends until a termination nucleotide is randomly incorporated. 5. The resulting DNA fragments are denatured into ssDNA. 6. The denatured fragments are separated by gel electrophoresis and the sequence is determine 1. 2. 3. 4. Procedure 1. Template preparation: 1) M13-forward-sequence copies of template strand to be sequenced must be preparedwith short known sequences at 3’ end of the template strand. 2) A DNA primer is essential to initiate replication of template, so primer preparation ofknown sequences at 3’end is always required. 3) For this purpose a single stranded cloning vector M13 is flanked with template strand at 3’end which serves as binding site for primer. 2. Generation of nested set of labelled fragments: 1) Copies of each template is divided into four batches and each batch is used for differentreplication reaction. 2) Copies of standard primer and DNA polymerase I are used in all four batches. 3) To synthesize fragments that terminates at A, ddATP is added to the reaction mixture on batch I along with dATP, dTTP,dCTP and dGTP, standard primer and DNA polymerase I. 4) Similarly, to generate, all fragments that terminates at C, G and T, the respective ddNTPs i.e. ddCTP, ddGTP and ddTTP are added respectively to different reaction mixture on different batch along with usual dNTPs. 3. GEL ANALYSIS & DETERMINATION OF DNA SEQUENCE The last step simply involves reading the gel to determine the sequence of the input DNA. Because DNA polymerase only synthesizes DNA in the 5’ to 3’ direction starting at a provided primer, each terminal ddNTP will correspond to a specific nucleotide in the original sequence (e.g., the shortest fragment must terminate at the first nucleotide from the 5’ end, the second-shortest fragment mustterminate at the second nucleotide from the 5’ end, etc.) Therefore, by reading the gel bands from smallest to largest, we can determine the 5’ to 3’ sequence of the original DNA strand. In manual Sanger sequencing, the user reads all four lanes of the gel at once, moving bottom to top, using the lane to determine the identity of the terminal ddNTP for each band. For example, if the bottom bandis found in the column corresponding to ddGTP, then the smallest PCRfragment terminates with ddGTP, and the first nucleotide from the 5’ end of the original sequence has a guanine (G) base. In automated Sanger sequencing, a computer reads each band of the capillary gel, in order, using fluorescence to call the identity of each terminal ddNTP. In short, a laser excites the fluorescent tags in each band, and a computer detects the resulting light emitted. Because each of the four ddNTPs is tagged with a different fluorescent label,the light emitted can be directly tied to the identity of the terminal ddNTP. The output is called a chromatogram, which shows the fluorescent peak of each nucleotide along the length of the template DNA. HOW TO READ SANGER SEQUENCING RESULTS Reading the Sanger sequencing results properly will depend on which of the two complementary DNA strands is of interest and what primer is available. If the two strands of DNA are A and B and strand A is ofinterest, but the primer is better for strand B, the output fragmentswill be identical to strand A. On the other hand, if strand A is of interest and the primer is better for strand A, then the output will be identical to strand B. Accordingly, the output must be converted back to strand A. So, if the sequence of interest reads “TACG” and the primer is best for that strand, the output will be “ATGC” and, therefore, must be converted back to “TACG”. However, if the primer is better for the complementary strand (“ATGC”), then the output will be “TACG”, which is the correct sequence. In short, before starting, you need to know what you’re targeting andhow you’re going to get there! So keeping this in mind, here is an example of the former example (TACG -> ATGC > TACG). If the dideoxynucleotides labels are T = yellow, A = pink, C = dark blue, and G = light blue, you will end up with the short sequences primer-A, primer-AT, primer-ATG, and primer-ATGC. Once the fragments havebeen separated by electrophoresis, the laser will read the fragments in order of length (pink, yellow, light blue, and dark blue) and produce a chromatogram. The computer will convert the letters, so the final sequence is the correct TACG. SANGER SEQUENCING VS. PCR Sanger sequencing and PCR use similar starting materials and can be used in conjunction with each other, but neither can replace the other.PCR is used to amplify DNA in its entirety. While fragments of varying lengths may be produced by accident (e.g., the DNA polymerase might fall off), the goal is to duplicate the entire DNA sequence. To that end, the “ingredients” are the target DNA, nucleotides, DNA primer, and DNA polymerase (specifically Taq polymerase, which can survive the high temperatures required in PCR). In contrast, the goal of Sanger sequencing is to generate every possible length of DNA up to the full length of the target DNA. That is why, in addition to the PCR starting materials, the dideoxynucleotides are necessary. Sanger sequencing and PCR can be brought together when generating the starting material for a Sanger sequencing protocol. PCR can be used to create many copies of the DNA that is to be sequenced. Having more than one template to work from makes the Sanger protocol more efficient. If the target sequence is 1,000 nucleotides long and there is only one copy of the template, it is going to take longer to generate the 1,000 tagged fragments. However, if there are several copies of the template, in theory it will take less time to generate all 1,000 of the tagged fragments. Applications Sanger sequencing gives high-quality sequence for relatively long stretches of DNA (up to about 900900900 base pairs). It's typicallyused to sequence individual pieces of DNA, such as bacterial plasmids or DNA copied in PCR. SNP and indel genotyping. Find common, known and unknown sequence variations. Study a gene or its part. 16 and 18S rRNA gene sequencing. DNA fingerprinting. Finds mutations and genes associated with a disease. Practical 13 Column chromatography What is column chromatography? Column chromatography is described as the useful technique in which the substances to be isolated are presented onto the highest point of a column loaded with an adsorbent (stationary phase), go through the column at various rates that rely upon the affinity of every substance for the adsorbent and the solvent or solvent mixture, and are typically gathered in solution as they pass from the column at various time. The two most common examples of stationary phases for column chromatography are silica gel and alumina while organic solvents are regarded as the most common mobile phases. Column Chromatography principle The main principle involved in column chromatography is the adsorption of the solutes of the solution with the help of a stationary phase and afterward separates the mixture into independent components. At the point when the mobile phase together with the mixture that requires to be isolated is brought in from the top of the column, the movement of the individual components of the mixture is at various rates. The components with lower adsorption and affinity to the stationary phase head out quicker when contrasted with the greater adsorption and affinity with the stationary phase. The components that move rapidly are taken out first through the components that move slowly are eluted out last. The adsorption of solute molecules to the column happens reversibly. The pace of the movement of the components is communicated as: Rf = the distance traveled by solute/ the distance traveled by the solvent Where Rf is called retardation factor Column Chromatography components Components of a typical chromatographic system using a gas or liquid mobile phase include: Stationary phase – Generally it is a solid material having a good adsorption property and should be suitable for the analytes to be separate. It should not cause any hindrance in the flow of the mobile phase. Mobile phase and delivery system – This phase is made up of solvents that complement the stationary phase. The mobile phase acts as a solvent, a developing agent (promotes separation of components in the sample to form bands), and an eluting agent (to remove the components from the column that are separated during the experiment). Column – For liquid chromatography: 2-50cm long and 4mm internal diameter, fabricated with stainless steel For gas chromatography: 1-3m long and 2-4mm internal diameter, fabricated either with glass or stainless steel A column’s material and its dimension are very crucial to support the stationary phase and promote effective separations. Injector system – Responsible for delivering test samples to the column’s top in a reproducible pattern. (Molecular Biology BIOT-304, BCH 305) Detector and Chart Recorder – This gives a continuous record of the presence of the analytes in the eluate as they come out from the column. Detection relies on the measurement of a physical parameter (like visible or UV adsorption). On the chart recorder, each separated analyte is represented by a peak. A collector at the bottom is placed at the bottom end of the column set up to collect the separated analytes. Column chromatography Procedure The steps included in the column chromatography are: Preparation of the column Mostly the column is comprised of a glass tube with an appropriate stationary phase The bottom end of the column is packed with a glass wool/cotton wool or an asbestos pad after which the stationary phase is packed. After packing the column, a paper disc is placed on the top to avoid the disturbance of the stationary phase during the introduction of the sample or mobile phase. The disturbance in the stationary phase (adsorbent layer) leads to the irregular bands of separation. Two types of preparing the column, known as packing techniques namely: Dry packing technique – The amount of absorbent needed is added as a fine dry powder in the column and the solvent flows freely through the column until equilibrium is achieved. Wet packing technique – The slurry of adsorbent is prepared along with the mobile phase and is poured into the column. It is regarded as the ideal technique for packaging. Page 38 of 41 (Molecular Biology BIOT-304, BCH 305) The column should be properly washed and completely dried before inuse. Introduction of the sample The sample (a mixture of components) is dissolved in the minimum amount of the mobile phase. At one instant, the sample is introduced into the column and on the top portion of the column, it is absorbed. Through the elution process, the individual sample can be isolated from this zone. Elution technique Through this technique, the individual components are separated completely from the column. The process of elution can be carried out by employing two techniques: Isocratic elution technique – Throughout the procedure, a solvent of the same polarity or same solvent composition is utilized. Example: Use of chloroform alone Gradient elution technique – Throughout the separation procedure, solvents of gradually increased polarity or increased elution strength are utilized. Example: Benzene → Chloroform → Ethyl acetate → Chloroform Detection of Components In case the mixture separated in a column chromatography procedure are colored compounds, then monitoring the separation progress is simple. In case the compounds undergoing separation are colorless, then small fractions of the eluent are sequentially collected in tubes that are labeled. Thorugh TLC, the composition of each fraction is determined. Column chromatography uses Column chromatography is one of the versatile methods for purifying and separating both solids and liquids. Major applications: To isolate active constituents To separate compound mixtures To remove impurities or carry purification process To isolate metabolites from biological fluids To estimate drugs in drug formulations or crude extracts Practical #14 Thin Layer chromatography Chromatography is an important biophysical technique that enables the separation, identification, and purification of the components of a mixture for qualitative and quantitative analysis. Page 39 of 41 (Molecular Biology BIOT-304, BCH 305) In this physical method of separation, the components to be separated are distributed between two phases, one of which is stationary (stationary phase) while the other (the mobile phase) moves in a definite direction. Depending upon the stationary phase and mobile phase chosen, they can be of different types. Thin Layer Chromatography (TLC): Thin Layer Chromatography can be defined as a method of separation or identification of a mixture of components into individual components by using finely divided adsorbent solid / (liquid) spread over a plate and liquid as a mobile phase. Components of Thin Layer Chromatography (TLC) TLC plates TLC chamber Mobile Phase A filter Paper Principle of Thin Layer Chromatography (TLC) TCL is based on the principle of separation through adsorption type. The separation relies on the relative empathy of compounds towards the mobile phase and stationary phase.The principle of TLC is the distribution of a compound between a solid fixed phase (the thin layer) applied to a glass or plastic plate and a liquid mobile phase (eluting solvent) that is moving over the solid phase. Procedure of Thin Layer Chromatography (TLC) The stationary phase is applied onto the plate uniformly and then allowed to dry and stabilize. These days, however, ready-made plates are more commonly used. 7. With a pencil, a thin mark is made at the bottom of the plate to apply the sample spots. 8. Then, samples solutions are applied on the spots marked on the line in equal distances. 9. The mobile phase is poured into the TLC chamber to a leveled few centimeters above the chamber bottom. 10. A moistened filter paper in mobile phase is placed on the inner wall of the chamber to maintain equal humidity (and also thereby avoids edge effect). 11. Now, the plate prepared with sample spotting is placed in TLC chamber so that the side of the plate with the sample line is facing the mobile phase. Then the chamber is closed with a lid. Page 40 of 41 (Molecular Biology BIOT-304, BCH 305) 12. The plate is then immersed, such that the sample spots are well above the level of mobile phase (but not immersed in the solvent) for development. 13. Sufficient time is given for the development of spots. 14. The plates are then removed and allowed to dry. 15. The sample spots are then seen in a suitable UV light chamber, or any other methods as recommended for the given sample. Retention Factor (Rf ) Value The behaviour of a compound on a TLC is usually described in terms of its relative mobility or Rf value. Rf or Retention factor is a unique value for each compound under the same conditions. The Rf for a compound is a constant from one experiment to the next only if the chromatography conditions below are also constant: solvent system adsorbent thickness of the adsorbent amount of material spotted temperature . Page 41 of 41