MANUAL OF INSTRUCTION GUARANTEE: For a period of one year f r o m date of installation, new e q u i p m e n t manufactured by UNIVERSAL X-RAY and sold and installed by an authorized UNIVERSAL X-RAY dealer is guaranteed as f o l l o w s : Any parts proving defective will be repaired or replaced free of c h a r g e , F.O.B. factory, if defective parts are returned to the f a c t o r y for inspection, c h a r g e s prepaid. The guarantee does not apply to glassware or high-voltage cables or to d a m a g e caused by accident, misuse, neglect, or normal wear and tear, and is void if service is p e r f o r m e d by persons other than authorized UNIVERSAL dealers or representatives. Components not manufactured by UNIVERSAL X-RAY are not c o v e r e d by the above, but carry the guarantee of the original manufacturer. U N I V E R S A L X-RAY reserves the right to pass j u d g m e n t on cause of breakage or failure. NOTE: This manual contains information for assemblers and users of a certified p r o d u c t as required by the Regulations f o r the Administration and Enforcement of the Radiation Control f o r Health and Safety A c t of 1968 issued by the U.S. Department of Health, Education and Welfare. This manual is to be delivered to the assembler of the product, w h o , upon c o m p l e t i o n of installation, is to transmit it to the p r o d u c t user. Any person or persons responsible for failure to c o m p l y with these instructions shall be liable for resultant d a m a g e s in a c c o r d a n c e w i t h the law. Careful attention to all the details and instructions enclosed are necessary for a p r o p e r installation. Familiarity with the requirements of the DHEW and experience with x-ray e q u i p m e n t , its installation, use, and maintenance are absolute prerequisites before attempting any w o r k with these units. Careful attention to the National Electrical c o d e is also r e q u i r e d . This e q u i p m e n t unless specifically indicated is not designed for use in the presence of explosive or inflamable gases, and s h o u l d not be installed in flamable a t m o s p h e r e s . S h o u l d there be any questions, p r o b l e m s , or suggestions either with these instructions or with the e q u i p m e n t supplied by Universal please write or call U n i v e r s a l X-Ray 4014 W. G r a n d Ave., C h i c a g o , Illinois 60651, 312/276-4487. We will w e l c o m e any a p p l i c a b l e c o m m e n t s or suggestions. CHEMETRON Universal X-Ray Division of Chemetron Corporation 4014 West Grand Avenue Chicago, Illinois 60651 U.S.A. Telephone: (312) 276-4487 . Cable UNIVAPROCO #040876 "LITTLE GIANT 30" a small x-ray unit doing a BIG JOB Universal X-Ray Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 ELECTRICAL ASSEMBLY, OPERATION, and MAINTENANCE OF X-RAY MACHINE Cat. No. 3205 "LITTLE GIANT" TUBEHEAD 30 mA, 95 kVp. INDEX section one Cat. No. 3335 "LITTLE GIANT" RADIOGRAPHIC CONTROL 30 mA., 95 kVp. page one GENERAL DESCRIPTION page t w o section two ASSEMBLY Dear User: Y o u r n e w UNIVERSAL x-ray m a c h i n e is a fine p r o d u c t r e p r e s e n t i n g over 40 years of e x p e r i e n c e in p r o d u c i n g simple to o p e r a t e a n d t r o u b l e free e q u i p m e n t . s e c t i o n three Read t h i s i n s t r u c t i o n m a n u a l c a r e f u l l y t o familiarize yourself w i t h the s i m p l e p c e d u r e s involved in p r o d u c i n g q u a l i t y r a d i o g r a p h s . Y o u w i l l find that every r a d i o g r a p h is a r e w a r d i n g e x p e r i e n c e f i l l e d w i t h interest a n d s a t i s f a c t i o n . MAINTENANCE page five OPERATION page eight section four s e c t i o n five page ten MISCELLANEOUS NOTE: This section contains special additions and corrections to assembly, operation and maintenance instructions and should be consulted before p r o c e e d i n g with assembly. We at UNIVERSAL are p r o u d of our p r o d u c t s and k n o w y o u r x-ray unit w i l l p r o v i d e many y e a r s of useful and e n j o y a b l e service. T h a n k y o u f o r s e l e c t i n g o u r equipment. This section provides a means for expanding the usefulness of this manual by insertion of newly developed data gained f r o m field experience and for recording data d e v e l o p e d by the assembler or user a p p l i c a b l e to local conditions. Universal X-Ray P r o d u c t s Inc. M e r t o n Moss p r e s . s e c t i o n six WIRING DIAGRAMS WIRING D I A G R A M S AND S C H E M A T I C S R a d i o g r a p h i c c o n t r o l Cat. No. 3335 T w o types of wiring information are s u p p l i e d . The Wiring Diagram generally s h o w s 'where the wires go,' and is useful in physically locating connections f o r servicing. The schematic diagram s h o w s the c i r c u i t in a manner which makes it easy to locate the source of trouble t h r o u g h the use of logic and d e d u c t i o n . Both types of d i a g r a m s are supplied for each c o n t r o l / t u b e h e a d combination c o v e r e d by this instruction. s e c t i o n one GENERAL DESCRIPTION The Cat. No. 3205 " L i t t l e Giant 3 0 " t u b e h e a d is d e s i g n e d for use with a variety of Universal x-ray units identified as f o l l o w s : Cat. No. 3205 " L i t t l e G i a n t " 30mA tubehead. Cat. No. 3 1 1 0 - 3 M o b i l e m a s t e r Cat. No. 3166-3 S c o p e m a s t e r f l u o r o s c o p e , Cat. No. 3180-3 Podiatry x-ray, Cat. No. 3175-3 Dentex dental x-ray, Cat. No. 3114 F i e l d m a s t e r portable x-ray for veterinary use, and the C e p h a l o m e t r i X 3 0 " used for lateral head and intra oral radiography. These applications utilizing the " L i t t l e Giant 30" generator and c o n t r o l , require the same general procedures d e s c r i b e d in this booklet under assembly maintenance and use. Special procedures if any are c o v e r e d in specific instructions for the model or a p p l i c a t i o n involved. page one Printed In U.S.A. oioa.74 " L I T T L E GIANT 3 0 " a small x-ray unit d o i n g a BIG JOB Universal X-Ray Power requirements section t w o ASSEMBLY IMPORTANT NOTICE The x-ray machine s u p p l i e d by Universal X-Ray P r o d u c t s Inc. and covered in this manual will p e r f o r m reliably when o p e r a t e d , maintained, and repaired in a c c o r d a n c e with the instructions of this manual. It should be c h e c k e d periodically and repaired as necessary to insure reliable operation. B r o k e n , missing, w o r n , inaccurate or non-functioning parts should be replaced. The e q u i p m e n t or its c o m p o n e n t s should not be m o d i f i e d . Universal X-Ray P r o d u c t s Inc. cannot assume responsibility for any malfunction of this equipment resulting f r o m improper operation, maintenance or repair, or if any of its c o m p o n e n t s are d a m a g e d or m o d i f i e d . Anyone using this x-ray machine must know and understand the dangers of excessive radiation exposure. This equipment is sold with the understanding that since its proper use and a p p l i c a t i o n is in the hands of the operator and b e y o n d the control of the manufacturer or his agents that Universal X-Ray P r o d . Inc. disclaims all responsibility for any injury resulting from i m p r o p e r use and a p p l i c a t i o n of this equipment. Unpacking The shipper is relieved of any responsibility for d a m a g e during s h i p m e n t after it is picked up by the carrier. Examine all cartons carefully at time of delivery. If d a m a g e is apparent have delivering driver write a " b a d o r d e r " note on all copies of the freight bill and sign it. Should you d i s c o v e r c o n c e a l e d damage, immediately notify the transportation agent and ask for an " I n s p e c t i o n Report of D a m a g e . " Carriers will not accept c o n cealed damage c l a i m s if filed after 15 days from date of receipt of merchandise. Open crates and c a r t o n s carefully and do not dispose of t h e m until you have located all parts and the m a c h i n e is fully assembled. Line v o l t a g e r e g u l a t i o n at a maximum rated line current must not be greater than 5 % . Percent line voltage r e g u l a t i o n = 100 (Vn Vn — V I ) where: Vn = No load line v o l t a g e . VI = Line voltage at m a x i m u m line current. A l t e r n a t e line v o l t a g e w h i c h will provide normal operation are indicated on the wiring d i a g r a m of the p a r t i c u l a r m o d e l . M a x i m u m line c u r r e n t at an alternate line voltage = ( V r ) ( A r ) where: Vr = Rated line v o l t a g e (no-load). A r = M a x i m u m line c u r r e n t when line voltage is equal to the rated line voltage (no-load). Va = Alternate line v o l t a g e to be used. A m p l e power is essential for the efficient operation of any x-ray machine. To allow your x-ray machine to be used at full capacity, the power line to w h i c h y o u r machine is connected must meet the minimum requirements listed below. If y o u r power line does not meet these requirements, you will have limited y o u r use of the x-ray machine to exposure values that co me within the limits of y o u r power line. Machine, " L i t t l e Giant" 30 mA Nom, line voltage, 240 volts single phase. Current rating, 15 amperes. Sw. box and fuse size, 15 amperes. Wire size f r o m Power Trans, to sw. box. 50 ft. #12, 100 ft. #10, 200 ft. # 8 . Machine, " L i t t l e Giant" 30 mA. Nom. line voltage, 120 volts single phase. Current rating, 30 amperes. Sw. box and fuse size, 30 amperes. Wire size f r o m Power trans, to sw. box. 50 ft. # 1 0 , 1 0 0 ft. #8, 200 ft. # 6 . Use time lag fuses. NOTE 1 Maximum daily line voltage variations due to causes other than the x-ray machine load, should not exceed ±21/2% of the nominal line voltage. NOTE: 2 These recommendations are based on an overall line voltage regulation not exceeding 5% measured at the x-ray control with the maximum rated output. NOTE 3 When operating at reduced line voltages more line current is required. For e x a m p l e , a 208 volt line requires proportionately more line current than a 240 volt line and may thus e x c e e d the capacity of the line to stay within the required Line Voltage Regulation. The wire size or other factors may limit the capacity of the line. Line Current is directly proportional to the p r o d u c t of kV and mA. and operation at a lesser p r o d u c t of kV. and mA. will result in a proportionately lower line current requirement. The mA. output can be limited to the required m a x i m u m in multiples of 5 mA. from 15 mA. to 30 mA. Use one value of mA. and establish the mA. presetting value, see Calibration of Filament Preset on page 4. See wiring diagram for setting x-ray filament adjustment (X-FIL ADJ) to a preselected m a x i m u m of mA. output. NOTE 4 M a x i m u m Use of Limited Power Line. Where limited power is available to operate all the equipment in a facility including the x-ray unit, it is possible to arrange for sharing the line. Certain heavy loads such as air conditioners, heaters and or sterilizers can be turned off for the short period of time required to take an x-ray exposure. Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 ASSEMBLY Universal x-ray m a c h i n e s are the simplest possible to assemble and install. Just f o l l o w the few easy steps given below. The control is s h i p p e d with a 12 foot line c o r d and the t u b e h e a d with an 8 foot or 15 foot head to c o n t r o l cable attached to . the t u b e h e a d and terminating in a 6 wire polarized Jones plug w h i c h plugs into the control. FOR FIXED I N S T A L L A T I O N S Most state regulations require that " t h e operator station at the control shall be behind a protective barrier, either in a s e p a rate room, in a p r o t e c t e d b o o t h , or behind a shield w h i c h will intercept the useful beam and any radiation w h i c h has been scattered only o n c e " and further that "a w i n d o w of lead equivalent glass equal to that required by the adjacent barrier, or a mirror system, shall be p r o v i d e d , large enough and so placed that the o p e r a t o r can see the patient without having to leave the protected area during exposure. Electrical power must be supplied to a switch and fuse b o x located in a position readily accessible f r o m the control l o c a t i o n . This is a National Electric Code requirement however, the National Electrical Code permits " e q u i p m e n t c o n n e c t e d to a 120 volt branch circuit of 30 a m p e r e s or less (to use) a g r o u n d i n g type cap and receptacle of proper r a t i n g — a s a d i s c o n n e c t i n g means" Thus 30 mA units may be p l u g g eedd into standard 120 volt wall outlets, if the power supplied to that outlet satisfy the requirements listed under the heading POWER R E Q U I R E M E N T S . Mobile R a d i o g r a p h i c e q u i p m e n t requires the o p e r a t o r to make an exposure at a safe distance and not in the direct b e a m . Check the local state requirements for safety regulations w h e n using mobile x-ray units. Handle M o b i l e X-ray Units carefully In order to keep the w e i g h t d o w n to a m i n i m u m , Universal mobile x-ray units were d e s i g n e d for careful handling by capable, authorized operators. They s h o u l d not be h a n d l e d carelessly or be b o u n c e d a r o u n d . M o b i l e x-ray units contain delicate scientific equipment. They are d e s i g n e d only for interior use on smooth level floors and unobstructed and level entrances and exits from rooms and elevators. The t u b e h e a d should be p l a c e d in the s e c u r e d position before porting. (continued on next page) page t w o R0A0776 R031777 "LITTLE GIANT 3 0 " a small x-ray unit d o i n g a BIG J O B Universal X-Ray Division of Chemetron Corporation 4014 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 section two ASSEMBLY Frequently c h e c k all locks, wheels, bolts, nuts, etc. and the entire unit to be sure nothing was d a m a g e d , knocked off or shaken loose d u r i n g moving and h a n d l i n g . DO NOT OPERATE E Q UIPM ENT until a c c e p t a b l e c o r r e c t i o n s are made. With frequent inspections and p r o p e r care this unit will provide continuous useful service. A T T A C H I N G T U B E H E A D TO SUPPORT STRUCTURE Attach the t u b e h e a d to its support by inserting the yoke pivot shaft into the support receptacle and securing by tightening the locking screw in the support. This applies to the 30 mA mobile x-ray unit Cat. No. 3110) and other Universal supporting structures w h i c h secure the fork in a horizontal position. When the fork of the t u b e h e a d is inserted in a s u p p o r t i n g structure receptacle so that it hangs vertically such as the Cat. No. 3180 Podiatry x-ray unit, then care should be taken so that the t w o special screws w h i c h secure the w a s h e r to the stud of the fork are w i r e d in with the wire p r o v i d e d . The two special screws have holes for this wire. See assembly instructions included with the supporting structure. ATTACHING COLLIMATOR TO T U B E H E A D A suitable c o l l i m a t o r must be used with the 30 mA t u b e h e a d s . T h e tubehead cover has a port thru w h i c h the primary x-ray beam is e m i t t e d , and s u r r o u n d i n g the port are f our holes for a t t a c h m e n t of the collimator. A t t a c h m e n t bolts are s u p p l i e d . Check them out f o r p r o p e r fit. "emove the cover from the tu b e h e a d by amoving the seven s c r e w s which attach the c o v e r to the t u b e h e a d , and lifting off the cover. Secure the c o l l i m a t o r to the cover with the bolts s u p p l i e d and tighten bolts well. Be sure c o l l i m a t o r front is in the same d i r e c t i o n on the c o v e r as the front of the t u b e h e a d . Be careful that the bolts used do not project too far inside the cover since this may cause p u n c t u r e of the expansion bag. C h e c k carefully and insert a fiber or rubber barrier if necessary to protect e x p a n s i o n bag. Put heads of screws inside c o v e r if possible. Replace cover on t u b e h e a d with cover attachment s c r e w s . ADJUSTING THE CONTROL TO LINE VOLTAGE HOW TO CHECK FOR CORRECT LINE V O L T A G E SETTING The control line c o r d is normally tagged indicating the line voltage at w h c h it is set. It usually is set at 230 volts. This reduces possibility of damage to the control which w o u l d o c c u r if the control were set at 120 volts and accidentally connected to a 230 volt line. If a v o l t m e t e r for checking the line is not available then the control kVp meter, w h i c h is a voltmeter can be used. This adjustment should be made with the line c o r d disconnected from the s o u r c e . For c h e c k i n g purposes the t u b e h e a d plug can be d i s c o n n e c t e d from the c o n t r o l . Check the setting inside the control to be sure it corresponds with the indication on the tag. If used on a different voltage than on the tag, change voltage on tag or remove tag. With the control selector knob pointing to " V " on the panel and the main switch in the " O N " position, the top scale (preset volt scale) on the voltmeter s h o u l d read a p p r o x i m a t e l y 117 volts. This 117 volt position on the meter is identified by the letter " V . " To check or reset the line voltage inside the control, disconnect the cable from the power outlet. Place the control upside down on a solid table using a folded towel or blanket for padding to protect the control panel. Remove the two screws at the lower end of the sides of the control and lift the back panel out at the lower end and pull down away from the pins at the upper end of the control housing. Exercise care so that the interior wiring is never disturbed or damaged. The control terminal board is marked with numbers 9 thru 14. Using the chart below (also chart on wiring diagram on inside of back of control), connect the leads marked Lj and L to the terminals on the terminal board in the control corresponding to the line voltage (power source). 2 E X A M P L E : For a 118 volt line connect L1 to terminal 1 and L to terminal 12. Choose the terminals corresponding to a voltage closest to the actual line voltage. 2 LINE VOLTS 97 104 110 117 123 130 205 211 217 224 230 237 243 250 CONNECT L1 to 1 1 1 1 1 1 0/S1 0/S1 0/S1 0/S1 0/S1 0/31 0/S1 0/S1 The c o l l i m a t o r normally requires 24 volts for the light bulb. C h e c k the instructions s u p p l i e d with the c o l l i m a t o r for details on mounting and voltage and wattage requirements. Universal controls Cat. No. 3335 provide a 24 volt, 200 watt source at terminals C1 and C2. Consult wiring diagram. Bring the wire in t h r u the bottom of the control alongside the e x p o s u r e pushbutton cable and c l a m p securely. page three CONNECT L2 to 9 10 11 12 13 14 7 8 9 10 11 12 13 14 HOW TO SET KILOVOLTS Always set kilovolts before you turn on x-rays,never turn the kilovolt selector knob while the exposure switch is closed and x-rays are on. Turn on the main switch. Determine the required exposure factors you are going to use. For e x a m p l e — s u p p o s e you are going to use 15 mA and 80 kVp. Turn the kilovolt selector until the kilovoltmeter needle points to 80 on the 15 mA scale. If you cannot set the needle exactly on 80, then set it slightly above 80 rather than below. Since each step of the kilovolt selector changes kilovoltage by a p p r o x i m a t e l y 3 kVp, you can always set kilovoltage sufficiently close to the desired value. If you use a value of mA not s h o w n on the kilovoltmeter then approximate the position between the higher and lower mA lines and thus you will come fairly close to the c o r r e c t kVp. Low p o w e r e d units are s o m e t i m e s o p e r a t e d on the same circuit with other e q u i p m e n t and may be subject to voltage d r o p . T h i s should be considered in establishing exposure values. "LITTLE GIANT 30" a small x-ray unit doing a BIG JOB Universal X-Ray Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 section t w o ASSEMBLY AUTO TRANSFORMER VOLTAGE TABLES The voltages indicated in the tables shown will only be obtained if the line voltage has one of the indicated discrete values, and the line voltage tap is properly set. Since line voltage will probably vary a volt or so from one of the indicated values, all readings will be off by the same percentage. Ep Volts is the voltage between 1 AT or X2 and the selector switch w h i c h goes thru the contactor to X 1 . These are the voltages delivered to the primary of the hi-tension transformer. The Ep VOLTS can be read on the top (red) scale of the kilovoltmeter. Obtain preset reading for all values of mA. shown on the voltmeter. Make a table showing preset values as f o l l o w s : MA. KV KNOB S E T T I N G M No Load 1 P r i m a r y Ep V o l t s 6 0 Load Pri. Volts 59 at 5 mA. kVp. (kilovolts) 45 N 0 65 71 65 70.5 52 56 P Q R S T U V W X 77 83 88 95 102 108 115 123 129 76 82 88 94.5 101 127 65 71 76 81 75 81 87 93 60 107.5 114 120.5 87 91 96 101 106 112 118.5 124.5 Load Pri. Volts 58 at 1 5 mA. k V p . ( k i l o v o l t s ) 40 63.5 44 50 55 60 64 69 74 80 85 92 Load Pri. Volts 57 at 2 0 mA. k V p . ( k i l o v o l t s ) 32 62.5 68.0 73.5 80 86 92 98 105 111 117 39 44 48 53 59 64 71 76 80 85 Load Pri. Volts 56.5 at 2 5 m A . k V p . ( k i l o v o l t s ) 28 62.5 68 73.5 79 85 91 98 104 109.5 116 Load Pri. Volts 56 at 3 0 mA. k V p . ( k i l o v o l t s ) 24 69 34 38 44 50 54 59 62 67.5 73 78.5 84 . 5 90.5 30 35 40 45 50 55 (continued M A X I M U M M I L L I A M P E R E LIMIT SET Remove the rear panel from the control for access to the X-Fil limiting resistor. (Section 4). With the milliampere selector fully clockwise, set control panel mA. selector knob to L position, move the band on this resistor to deliver about 32 to 33 mA. at 60 kVp. and tighten band locking screw. On the Cat. No. 3334 Fluoroscopic only control, position the band to restrict the fluoroscopic mA. to the maximum legal limit with the milliampere selector set fully clockwise. Under the Federal Performance Standards " F l u o r o s c o p i c equipment which is not provided with automatic exposure rate control, shall not be operable at any combination of tube potential and current, which will result in an exposure rate in excess of 5 roentgens per minute at the point where the center of the useful beam enters the patient." etc. etc. Measurements should be made 1 c M . away from the fluoroscopic panel thru which the radiation passes before entering the patient or subject. CALIBRATION OF THE FILAMENT PRESET Since the milliammeter only reads DURING an exposure, it is necessary to provide the operator with a preset calibration so that he may preset milliamperes before the exposure begins. Once these preset values are obtained it is not necessary to make test exposures before every radiograph is taken, in order to set the control to deliver the desired value of milliamperes. Periodic checks should be made to insure that some degree of a c c u r a c y is maintained. Two different methods are simultaneously available to preset milliamperes. a. Reading the position of the milliampere selector knob and, b. pressing down the "Filament Preset S w i t c h " and reading the "Preset V o l t s " scale on the voUmeter. 99.5 65 96.5 60 from 71 76 80 103 109 115.5 72 76 65 preceding Milliampere Selector Preset Volts 96 123 91 122 85 121.5 81 column) For accuracy of mA. presetting the preset voltmeter reading method should be used. Here is how to calibrate the filament preset. 1. Turn on the main switch on control and set kilovoltage control knob for an average value of kVp. that will be used most frequently (See sheet on exposure technique). About 80 pKv. is a g o o d average setting. 2. Assume that a reset value for 15 milliamperes is to be obtained. Set the timer for approximately two seconds and make an exposure using one hand on the milliampere selector knob to gradually (but quickly) bring up the reading on the mA. meter to the desired 15 mA. Care must be exercised to make these presettings accurate. When the timer button is pressed during an exposure, watch the milliampere meter needle. If it falls below the requirements this may be compensated for q u i c k l y by advancing the control during the exposure. Some corrections of preset values may have to be made from time to time because of changing filament characteristics. Some variation in milliamperes can also be caused by line fluctuations during exposure and cannot be c o r r e c t e d by changing preset values. FOR REVISED PANEL LAYOUT INTRODUCED APRIL, 1979 SEE PAGE 1 1 . HOW TO A L I G N THE C O L L I M A T O R It is a legal requirement that the collimator light beam be aligned with the x-ray beam so that misalignment in either direction does not e x c e e d 2% of the source image receptor distance (SID). For instance, if the source image receptor distance (SID) is 37 /2 inches, the maximum allowable misalignment is 2% of 37 /2 = 3/4 inches. 1 1 NOTE: Exposures under 7 / 1 0 seconds will not permit the milliampere needle to reach the c o r r e c t value of mA. Be sure test readings are taken at the point w h e r e the needle comes to a momentary rest. When making an exposure below 7 / 1 0 seconds the needle will swing past or fall short of the actual mA. passing thru the tube and therefore it is necessary to use the preestablished setting of the mA. control knob. Since the c o l l i m a t o r is packed separately it must be aligned at installation. Refer to the instruction sheet packed with the c o l l i m a t o r for location of adjustments and information on how to make the adjustments. C A U T I O N : Do not exceed tube rating chart in making test exposures. Be sure that tube port is not directed at operator or loaded cassette or film since x-rays are being p r o d u c e d during testing. Hold down the "filament p r e s e t " switch and accurately read and record the preset volts reading on the top scale of the pKv. meter. Record the position of the milliampere selector knob (Example A5 or B6.5 etc.) In o r d e r to again obtain this value of milliamperes it will not be necessary to make an exposure. Merely set the milliampere selector knob to this predetermined position on the voltmeter. (continued page four on next page) "LITTLE GIANT 3 0 " a small x-ray unit d o i n g a BIG JOB Universal X-Ray Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 s e c t i o n three OPERATION Protective Measures X-ray e q u i p m e n t i m p r o p e r l y used may cause injury. All persons using this equipment must be familiar with the hazards involved and the necessary protective measures before using the e q u i p m e n t . This equipment must be operated only by informed c o m petent personnel. Even tho this equipment has built in p r o t e c t i o n against all radiation except the useful b e a m , no practical design can provide c o m p l e t e protection or c o m p e l the o p e r a t o r or his assistant to take adequate precautions to prevent the possibility of authorized or unauthorized persons, carelessly, unwisely, or unknowingly exposing themselves or others to direct or s e c o n d a r y radiation. Handbooks n Radiation Protection have been c o m p i l e d by the National C o m m i t t e e on Radiation P r o t e c t i o n and are available from NCRP Publications, B o x 4867, Washington, D.C. 20008. The o p e r a t o r must become familiar with and use this e q u i p m e n t as recomm e n d e d in these h a n d b o o k s . M A I N LINE SWITCH This equipment is sold with the understanding that since Universal X-Ray Products Inc. has no control over the way the equipment is used, Universal X-Ray P r o d u c t s Inc., its agents or representative cannot assume responsibility for injuries suffered by anyone thru the use of this equipment. All personnel must be protected from the useful beam and also f r o m the secondary x-rays that are produced whenever x-ray strikes any project. Various protective materials and devices are available. It is urged that such materials or devices be used. The main line switch is a c o m b i n a t i o n line switch and overload breaker. It is " T r i p F r e e , " that is, you cannot hold it c l o s e d if there is a short circuit or o v e r l o a d . LINE ON Pilot Light Ruby light indicates that the main switch is in the ON position. No e x p o s u r e can be made unless the key lock e x p o s u r e switch is in the ON position. The key cannot be removed in this position. The key lock exposure switch will not turn off the ruby light. It will only go off w h e n the main switch is in the OFF position. To keep this equipment properly calibrated, it must be electrically and mechanically inspected and c h e c k e d in a c c o r d a n c e with Maintenance Procedures listed in Section 4 of this instruction manual, at regular intervals of 1 year or less, depending on the age and amount of use the equipment receives. If trouble o c c u r s — c a l l your serviceman and do not operate the equipment until the repairs are c o m p l e t e d and c h e c k e d out. CHART FOR OBTAINING THE ACTUAL VALUES OF KILOVOLTAGE UNDER LOAD AT A PARTICULAR MILLIAMPERE (MA.) SETTING. KV KNOB SETTING N o L o a dd P r i m a r y Load at kVp. 5 mA. Lead Pri. Amps. Pri. Load Pri. at kVp. 20 mA. Lead Load at kVp. 25 mA. Lead Load at kVp. 30 mA. Lead Volts (kilovolts) Load at kVp. 15 mA. Lead N 0 P 60 65 71 77 83 59 65 70.5 76 82 45 52 56 2.9 2.9 2.9 M Volts Volts 60 Q 65 S R 88 94.5 95 94.5 71 76 3.0 3.0 3.0 3.1 T U 102 108 115 123 129 101 107.5 114 120.5 127 V w X 81 87 91 96 101 3.1 3.1 3.1 3.1 3.2 106 112 118.5 124 .5 80 85 58 63.5 69 75 81 87 93 99.5 (kilovolts) 40 44 50 55 60 64 69 74 Amps. 8.7 9.0 9.2 9.0 9.0 9.0 9.0 9.0 9.0 9.2 9.2 9.5 57 62.5 68.0 73.5 80 86 92 98 105 111 117 123 (kilovolts) 32 39 44 48 53 59 64 71 76 80 85 91 Amps. 11.6 11.6 11.6 11.5 11.6 11.6 11.6 11.6 11.6 11.6 11.6 11.6 56.5 62.5 68 73.5 79 85 91 98 104 109.5 116 122 X1 Pri. Volts Volts 92 96 (kilovolts) 28 34 38 44 50 54 59 65 71 76 80 85 Amps. 14.2 14.2 14.1 14.2 14.2 14.2 14.2 14.1 14.1 14.1 14.1 14.2 115.5 121.5 X1 Pri. Volts 56 62 67.5 73 78.5 84.5 90.5 96.5 103 109 (kilovolts) 24 30 35 40 45 50 55 60 65 72 76 81 Amps. 17 17 17 17 16.8 16.8 16.8 16.9 17 16.5 16.9 16.8 X1 This chart provides the actual draw in the primary of the high tension transformer lead at the various settings of milliamperes. This primary current will vary directly with the milliamperes and therefore be sure to check the accuracy of the mA. meter reading to establish the validity of any current measurements. The X1 current can be obtained using a clip o- ammeter. It will detect a short in the primary or secondary coils or a defective inverse suppressor. (See page 10 on checking the inverse suppressor.) page five 010274 REV. 5-20-75 8-7-76 "LITTLE GIANT 3 0 " a small unit doing a BIG J O B Universal X-Ray Division of Chemetron Corporation 4014 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 section three OPERATION READY PILOT HOW TO OPERATE THE M A C H I N E THE M I L L I A M M E T E R indicates that the line switch is turned on and that the machine is ready for operation. Universal X-ray machines are d e s i g n e d to be extremely simple to operate and all you have to do is to follow the few simple steps given below. However, before y o u operate this or any other x-ray machine, you must b e c o m e familiar with all of the necessary protective measures and also R a d i o g r a p h i c Technique. A single scale meter w h i c h reads directly in milliamperes. It indicates only during the exposure when current is flowing thru the x-ray tube and thus also serves as an "X-Ray O n " indicator. HOW TO SET TIME The timer on the Cat. No. 3335 control is a Radiographic timer with a time range to 6 seconds in 21 steps, and in approximate geometric progression. Rotate the timer selector knob to bring the desired time value under the pointer. An exposure thru the timer is made by depressing the exposure switch on the control and holding it d o w n until the timer turns off the exposure. Releasing the button will instantly terminate the exposure irrespective of the timer setting. Determine the technical factors (kV., mA. and time) you are going to use, then p r o c e e d as f o l l o w s : HOW TO TURN ON THE M A C H I N E Push the LINE SWITCH lever to the " O N " position. The kilovoltmeter will i m m e d i a t e l y indicate and the READY PILOT will light. HOW TO USE THE C O L L I M A T O R REMOTE EXPOSURE SWITCH A single c i r c u i t 2 wire Cat. No. A979 hand switch or Cat. No. 3508 foot switch can be connected to terminals S1 and S2. The c o l l i m a t o r contains a lamp and m i r r o r system w h i c h : THE KILOVOLT SELECTOR b. Casts s h a d o w s of cross-lines s h o w i n g the center line of the x-ray b e a m in t w o directions. The kilovolt selector is a 12 position switch. Turn the selector knobs to the right (clockwise) to increase kilovoltage. Never t u r n the kilovolt selector d u r i n g an x-ray e x p o s u r e . The kilovolt selector must always be adjusted to the desired kilovolts before starting the x-ray e x p o s u r e . a. Projects a light beam covering the same area as the x-ray beam. This permits easier and more a c c u r a t e positioning of the patient and at the same time permits adjustment of the x-ray beam size to the size of the film being used, reducing both secondary radiation and patient dosage. THE KEY LOCK EXPOSURE SWITCH. The key lock exposure switch on the panel of the control prevents an exposure to be made when the key is removed from the switch. It does not prevent the main s w i t c h f r o m being left o n . Be sure the MAIN SWITCH is turned off when the x-ray machine is not in use. Adjust the c o l l i m a t o r to deliver an x-ray beam the same size as the film being used and at the distance being used. Use the c o l l i m a t o r light beam to position the patient so that the desired area is e x p o s e d . DO NOT leave the light beam on any longer than is necessary because the high intensity lamp has a short operating life. An audible signal in the f o r m of a single " b o n g " of a bell in the control will indicate the termination of an exp o su r e . Check instructions supplied with the c o l l i m a t o r for any more details. THE KILOVOLTMETER A l w a y s set kilovolts before you turn on x-rays never turn the kilovolt selector knob while the exposure switch is c l o s e d and x-rays are on. The kilovoltmeter is an ac voltmeter, calibrated in kilovolts. It also has a line marked 'V' w h i c h is used to c h e c k for proper line voltage adjustment. See kVp chart page four HOW TO SET KILOVOLTS Turn on the main line switch. Consult the exposure technique chart to determine the required exposure factors. For e x a m p l e : If the t e c h n i q u e chart requires 15 mA. and 80 kVp., turn the kilovolt selector until the kilovolt meter needle points to 80 on the 15 mA. scale. If you cannot set the needle exactly on 80, set it slightly above 80 rather than below. Since each step of the kilovolt selector changes kilovoltage by a p p r o x i m a t e l y 3 kVp., you can always set kilovoltage sufficiently close to the d e s i r e d value. page six The o p e r a t o r s h o u l d always observe the milliameter d u r i n g exposure in order to be sure that a n o r m a l exposure has been made. Observing this meter during exposure makes it possible for the o p e r a t o r to act quickly in turning off the main switch in the event failure of some c o m p o n e n t should o c c u r due to age or line conditions. Failure may manifest itself by a continuation of the exposure beyond the normal estimated time setting, or a wide fluxuation in the mA meter reading. On very short fractional exposures the mA. meter reading may fall short or swing past the actual mA setting. See the paragraph " H o w to make an x-ray e x p o s u r e " for more details. THE M I L L I A M P E R E SELECTOR The Milliampere Selector is directly below the Milliammeter. This is a stepless control. Turn the selector knob to the right to increase milliamperes but remember that the milliammeter only reads d u r i n g an e x p o s u r e (unlike the kilovoltmeter w h i c h reads all the time that the line switch is on). THE PRESET SWITCH Press the preset switch lever down to connect the kilovoltmeter to read x-ray filament t r a n s f o r m e r primary volts on the top meter scale (preset volts scale). Release the preset switch lever and it automatically reconnects the kilovoltmeter to indicate kilovolts. HOW TO SET M I L L I A M P E R E S Milliamperes can be preset by turning the miiliampere selector to a p r e d e t e r m i n e d position on its scale by using the kilovoltmeter preset scale. If you have an unusually stable power line, you might be able to use the miiliampere selector scale to preset milliamperes, but Universal X-Ray Products, Inc., recommends using the kilovoltmeter preset scale. To preset m i l l i a m p e r e s to the desired value by this m e t h o d , p r o c e e d as f ol low s : Refer to the table of preset volts versus mA. prepared when the machine was first installed (See Section 2 Calibration of the Filament Preset) and determine the preset volts c o r r e s p o n d i n g to the value of mA. you wish to use. Press d o w n on the preset switch handle, and turn the Miiliampere Selector Knob until the Kilovoltmeter needle points to the desired value of preset volts. Release the preset switch handle. "LITTLE GIANT 3 0 " a small unit doing a BIG JOB Universal X-Ray Division of Chemetron Corporation 4014 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 section three OPERATION HOW TO M A K E AN X-RAY EXPOSURE C A U T I O N : The milliampere meter is an indicator of x-ray being p r o d u c e d . If it should continue to read b e y o n d the time set on the timer, then quickly turn off the main switch and do not operate the e q u i p m e n t until c h e c k e d out and c o r r e c t e d . Set the timer to the desired value of time. Press the exposure button on the control. An approved handswitch or footswitch can be connected to terminals S1 and S2 in the control to also operate the exposure circuit. Observe the m i l l i a m m e t e r during the exposure. The milliammeter should indicate the value of mA. used, except that for short exposures, the needle may not have time to swing up and read the c o r r e c t value. On very short e x p o s u r e s the needle may swing past or fall short of the actual mA. setting. On exposures of about 1/2 second and higher the needle may swing past the -tual setting and momentarily come back rest at the actual mA. before dropping back to zero at the end of the exposure. This needle response is not to be construed as an indication of the actual mA. being higher or lower or c o m i n g up slowly. It is only the inertia of the meter that causes the needle to swing past the actual value. HOW TO TURN OFF THE M A C H I N E When you are thru using the machine push the LINE SWITCH lever to the ' o f f position. Never leave the control switch in the on position when not in use since this will shorten the life of the x-ray tube. N O T E : The exposure lock switch does not turn off the machine. It only locks the exposure circuit so that no exposure can be made when the key is removed. P O D I A T R Y - O T H E R APPLICATIONS Podiatry and other radiographic or fluoros c o p i c a p p l i c a t i o n s utilizing the " L i t t l e Giant 3 0 " generator and c o n t r o l , require the same general p r o c e d u r e s described in this b o o k l e t under assembly maintenance and use. Special p r o c e d u r e s if any are covered in specific instructions for the model or a p p l i c a t i o n involved. USE OF GRIDS WITH " L I T T L E GIANT 3 0 " On heavier, denser subject matter the use of a g r i d (by placing the cassette in the cassette tray in the grid cabinet) is desirable in o r d e r to reduce the amount of scatter radiation w h i c h tends to fog the film. An 8:1 ratio grid will p r o d u c e the best results. The " L i t t l e Giant 3 0 " generator will p r o d u c e a m a x i m u m of 30 milliamperes. On ordinary 115 volt lines (plug in receptacles) the m a x i m u m obtainable may be limited to a p p r o x i m a t e l y 15 or 20 milliamperes. page seven SUGGESTED TECHNIQUE FOR 30 M A . UNITS V e r y little difficulty in t a k i n g consistently g o o d r a d i o g r a p h s w i l l be e x p e r i e n c e d if the user w i l l observe some s i m p l e rules and u s e the f o l l o w i n g suggested techniques as a starting point. Variations to suit i n d i v i d u a l r e q u i r e m e n t s may be m a d e . DARKROOM 1. Use fresh solution. Change r e g u l a r l y based on the s i z e of tank, n u m b e r of films processed, and age of solut i o n . For average office using 5 g a l l o n tank change solution e v e r y three to f o u r w e e k s . 2. Use fresh f i l m , p r o p e r l y s t o r e d , and h a n d l e d w i t h p r o p e r filter and r e c o m m e n d e d b u l b size in d a r k r o o m l a m p . Keep film as far a w a y f r o m d a r k r o o m lamp as possible. CASSETTE SCREENS Use g o o d cassette and screens. Exposures g i v e n are based on the use of high speed screens. DISTANCE OVER FILM Use 40" distance f r o m top of table to tube ( w h i t e dot on f r o n t of tube) for ail over table and bucky or g r i d cabinet r a d i o g r a p h s , A convenient m e t h o d is to put a mark on the vertical c o l u m n , (a grease pencil line or a piece of tape) Use 7 2 " distance for chests and lateral cervicals. Measure f r o m the cassette holder to the t u b e , put a mark on the table i d e n t i f y i n g this position. Variations in subject density can be corrected by v a r y i n g the e x p o s u r e t i m e (staying w i t h i n tube ratings). If films are too dark reduce t i m e , if too light increase time. This assumes that f i l m d e v e l o p i n g t i m e remains fixed and is v a r i e d o n l y to compensate for variations in t e m p e r a t u r e of solution. EXTREMITIES - hands, w r i s t , ankle, e l b o w , etc. MILLIAMPERES 15 m a . KILOVOLTAGE " U " setting for all exposures (75 t o 8 0 pkv.) TIME 1 second using nonscreen f i l m in paper f o l d e r or nonscreen m o n o p a c k , five m i n u t e d e v e l o p i n g r e q u i r e d w i t h nonscreen f i l m . For Regular f i l m in cassette w i t h Universal X-Ray high speed second film u s e a p p r o x i m a t e l y 1/10 to 1/2 CHEST DISTANCE - 7 2 " CASSETTE w i t h screens MILLIAMPERES — 15 selector k n o b ma. Record p o s i t i o n of milliampere KILOVOLTAGE - " U " setting ( a p p r o x i m a t e l y 8 0 p k v . ) TIME - 1 / 2 t o 7 / 1 0 second CERVICAL SPINE Lateral o r P.A... u s i n g 7 2 " d i s t a n c e w i t h 8 x 1 0 c a s s e t t e i n adjustable w a l l cassette holder. MILLIAMPERES 15 M a . KILOVOLTAGE - " U " setting ( a p p r o x i m a t e l y 8 0 pkv.) TIME - 1 second SKULL, SINUS, CERVICAL SPINE, KNEES, SHOULDER CASSETTE w i t h screens KILOVOLTAGE - " U " setting ( a p p r o x i m a t e l y 7 5 t o 8 0 p k v . ) MILLIAMPERES — 1 5 ma. R e c o r d p o s i t i o n o f m i l l i a m p e r e selector k n o b TIME a p p r o x i m a t e l y 2 seconds (knee a n d shoulder 1 / 2 s e c o n d , s h o u l d e r t h r u b o d y 2 second) Time g i v e n b a s e d o n use w i t h 8 0 l i n e 8 : 1 f i x e d g r i d i n g r i d cabinet. Use a p p r o x i m a t e l y 3 5 % m o r e t i m e f o r exposures w i t h b u c k y . The stationary g r i d in a g r i d cabinet is r e c o m m e n d e d . PELVIS, HIP, LUMBAR SPINE, GASTRO INTESTINAL MILLIAMPERES — knob 1 5 ma. R e c o r d p o s i t i o n o f m a . selector KILOVOLTAGE - " U " setting ( 7 5 t o 8 0 p k v . ) TIME — A.P. V i e w — 3 seconds WITHOUT GRID - 1 - 1 / 2 seconds Time g i v e n based o n use w i t h 8 0 line 8 : 1 f i x e d g r i d i n g r i d cabinet. Use a p p r o x i m a t e l y 3 5 % m o r e t i m e f o r exposures w i t h b u c k y . The stationary g r i d in a g r i d cabinet is r e c o m m e n d e d . 4014 West Grand Avenue Chicago. Illinois 60651 U.S.A. Universal X-Ray "LITTLE GIANT 30" a small unit doing a BIG JOB Division of Chemetron Corporation 4014 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 section four section three MAINTENANCE OPERATION FOCAL SPOT RATINGS Note that these ratings apply to a cold tube. Never apply more than 7 0 % of these ratings after the tube has been heated up thru use. Effective focal spot size; 2.0mm. Effective target angle; 20 degrees TROUBLE SHOOTING The kilovolt meter in the control is an ac voltmeter and the preset scale is calibrated in ac volts. If you have no other meter available, in an emergency you can disconnect and tape up the leads going to the kilovoltmeter and use the kilovoltmeter for trouble shooting. Be sure to reconnect the leads when you are finished. NO KILOVOLTMETER READING If there is no kilovoltmeter reading when the main switch is " O N " make the following checks: 1. Check line supply. Disconnect switch may be open or fuses blown. If line supply is OK. ANODE HEAT STORAGE RATINGS Max. Cooling rate; 350 H.U./sec. Max. Anode Heat Storage capacity: 70,000 H.U. The anode heat storage rating of x-ray tubes used in the Cat. No. 3205 30 mA. tubeheads is 70,000 heat units. To determine the number of heat units you will put into the anode, multiply kVp. X mA. X seconds. Thus technique w h i c h calls for 80 kVp., 30 mA. and 1 1/2 econds, produces 80 X 30 X 1 1/2 = 3 6 0 0 heat units. Fluoroscoping at 80 kVp. and 5 mA. produces 80 X 5 X 60 seconds = 24,000 heat units a minute. TUBEHEAD HEAT STORAGE CAPACITY HEATING—With x-ray apparatus, most of the electrical energy developed in the high tension transformer and delivered to the tube is converted into heat. This conversion takes place at the focal spot on the target. The quantity of heat that is generated is proportional to the mA., and the pKv. and the time of exposure. At 15 mA., heat is generated at a m u c h higher rate than is conducted away f r o m the focal spot. It is therefore, important to limit the length of exposures at the higher mA. values within the limits of the tube rating chart. There is also the general temperature of the unit to be considered. After the unit has been run for a series of exposures, it may be found that the x-ray head is warm. The transformer, tube and oil will be warmer than the x-ray head. It is safe to keep using the equipment when the head is slightly warm, but extreme care must be taken not to exceed the limits per exposure for radiography either when the x-ray head is warm o r c o o l . Misuse of the x-ray tube through exceeding the safe exposure limits may melt the target and, of course, ruin the tube. 30 mA. TUBEHEAD OIL HEAT STORAGE CAPACITY 1. Oil capacity—approximately 5620 milliliters at 20ºC. 2. Check to see if line cable is properly plugged in (or line disconnect switch closed). 3. Be sure kilovolt selector is not set between contacts but is properly indexed on a contact point. 2. Oil expansion bag at 20°C is normally filled with approximately 100 milliliters of oil. 4. Check kilovoltmeter so that it is not open circuited or stuck. 3. Expansion bag total capacity (fully expanded) approximately 400 milliliters. NO MILLIAMMETER READING 4. Raising oil temperature in tubehead from 2 0 ° C to 95°C will add 300 milliliters of oil to expansion bag, thus making a total of 400 milliliters oil (maximum it will hold). 5. Lowering the oil temperature from 20°C to-5°C will empty bag. Circulated maximum operating temperature — 5 ° C t o 95°C. Suggested maximum operating temperatures + 5°C to 8 0 C ( 4 1 ° F t o 170°F). C DUTY CYCLE There is no cycle limitation on any of the control units listed herein. The tubehead duty cycle is determined for high energy short time exposures by the ratings of the x-ray tube and for low energy long time operation by the oil expansion capacity of the tubehead. TUBEHEAD RATINGS M a x i m u m rated peak tube potential 95kVp. M a x i m u m mA rating 30 milliameters. Leakage technique factors 95kVp, 2 mA. M i n i m u m inherent filtration with no added filter is 1.5 mm. aluminum at 90 kVp. page eight No exposure can be made unless the key lock exposure switch is in the ON position. The key cannot be removed in this position. X-Rays are turned on by timer or f o o t s w i t c h . 1. See if x-ray tube filament lights up. Do not make any exposure during this test. The tube is visible thru clear plastic window. Tubehead covers will need to be removed. Another check is to remove cover and put a voltmeter between terminals F1 and F2. If no voltage reading is obtained the trouble is possibly in head-to-control cable, control or line. If there is a voltage reading, the trouble is in t u b e h e a d , possibly a burned out filament. 2. Make sure the timer or footswitch is operating properly. This can be established by removing the head cover and checking to see if a voltage is obtained by means of a voltmeter or a test lamp connected between leads X1 and X2. If no voltage is obtained c h e c k back, starting with the timer (see wiring diagram). Be sure to cover port of tube and cover back on tubehead so as not to expose yourself or anyone else to radiation when making this test. The voltage between X1 and X2 ranges from approximately 60 to 120 volts. The actual voltage will be based on the position of the kilovoltage selector knob. The voltage reading can only be obtained when the exposure switch or foot switch is depressed. Cover tube port opening with lead in the event the machine is working and x-rays are being produced. "LITTLE G I A N T 3 0 " a small unit doing a BIG J O B Universal X-Ray Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 section f o u r MAINTENANCE 3. Disconnect terminal M1 and connect D.C. Milliameter (reverse lead if meter reads b a c k w a r d ) between this te r m in a l and the binding post f r o m w h i c h it w a s removed. If on making an e x p o s u r e t h r u the t i m e r this meter reads, then there is a short between M1 and Ground in the control or head to control c a b l e — o r the mA. meter on the control is shorted and needs replacement. 4. Determine if x-rays are being p r o d u c e d by means of a f l u o r o s c o p i c screen or by directing rays onto an open cassette in a darkened room to see if intensifying screen will g l o w w h e n x-rays are t u r n e d on. Exercise care by limiting size and direction of beam so as not to strike any personal with direct radiation. If x-rays are p r o d u c e d and milliameter still d o e s not r e a d — t h e n the possible t r o u b l e is a short between M1 and G r o u n d in the x-ray head. 5. If all of the above points c h e c k t h r u satisfactorily, then the t r o u b l e is in the x-ray head and is p r o b a b l y an open circuit. This can only be established by examining the inside of the x-ray h e a d . SERVICING THE T U B E H E A D Universal t u b e h e a d s are assembled by a careful and t h o r o u g h p r o c e s s w h i c h finally includes v a c u u m degassing of the t u b e h e a d and the insulating oil and filling the t u b e h e a d with oil under v a c u u m . This t h o r o u g h assembly process largely a c c o u n t s f o r the success and long o p e r a t i n g life of Universal t u b e h e a d s . At the same time, the need f o r this processing d u r i n g assembly makes it impractical for any but the most c o m p l e t e l y e q u i p p e d service stations to p e r f o r m repairs on Universal t u b e h e a d s . If a ehead is low in oil but otherwise sfactory, oil can be a d d e d by f o l l o w i n g the p r o c e d u r e in subsequent paragraphs. M I L L I A M M E T E R INDICATES BUT FLUCTUATES Check for l o w o i l c o n d i t i o n . The 30 mA. controls covered by this instruction manual have an adjustable x-filament limiting resistor mounted b e h i n d the meters and easily accessible for adjustment. This is a 50 watt, 125 o h m . cylindrical resistor, and adjustment is achieved by loosening the band locking screws and repositioning the b a n d . Retighten the band locking screw after adjusting. The bag should be only about 1/3 full (with no air) when head is at room temperature. Unless p r o p e r facilities are available this should be done at the factory. 2. This may be caused by a defective x-ray tube, either gassy or punctured or broken. 2. Loose connections within the unit or within the supply, or at the line c o r d plug may cause milliampere fluctuations. Loose connections of the filament circuit controlling the milliampere output can be c h e c k e d by examining the termination of F1 and F2 at the tubehead and flexing the c o r d where it enters the x-ray head to detect possible intermittent breaks in the wire. 3. Severe fluctuations of the milliameter may indicate a gassy x-ray tube. If all other points have been c h e c k e d , it can be assumed that the t r o u b l e is in the x-ray head and may be a gassy or broken tube. If just received retain shipping cartons and request inspection by carrier so as to file claim for c o n c e a l e d damages. 4. Severe fluctuations of the milliamperes may also indicate an a r c i n g in the x - r a y h e a d caused by an air pocket, or a possible short. See item 1 above (low oil). SERVICING THE C O L L I M A T O R The light bulb in the c o l l i m a t o r has a limited life, and will have to be replaced relatively frequently. Refer to instructions s h i p p e d with the c o l l i m a t o r for the specifications of replacement light bulbs, and for detailed instructions on how to replace the bulb. 1. Tip t u b e h e a d so the terminals (low voltage leads) are at highest point. This will c h e c k out f o r arcing or mA fluctuations c a u s e d by a low in oil c o n d i t i o n . In this position the oil will cover all the high voltage c o m p o n e n t s QUICK CHECK DATA Line c u r r e n t is d i r e c t l y p r o p o r t i o n a l to the p r o d u c t of kVp. a n d m A . and operation at a lesser p r o d u c t of k V p . and mA. will result in a p r o p o r t i o n a l l y " r e d u c e d line c u r r e n t . c l i p on a m m e t e r placed around one of the t w o line wires. T h e readinqs shown are f o r 230 volt lines. On 115 volt lines d o u b l e the value s h o w n in c h a r t . Reading line current draw of the control c a n q u i c k l y establish if the x-ray control and t u b e h e a d are f u n c t i o n i n g properly. Use a For precise evaluation of t u b e h e a d only see p a g e 5. Chart on page 5 provides normal current values measured in the X1 lead (primary of H.T.) of the t u b e h e a d . LINE CURRENT FOR 230 V O L T LINE (Approx.) X-Fil Preread Set kV C o n t r o l To Q 5 mA. 15 m A . 30 m A . 1.2 a m p s . 3.5 a m p s . 6.7 a m p s . X-FIL LIMITING R E S I S T O R - 3 0 mA. CONTROLS ONLY and any air pocket will be confined to this low voltage position of the tubehead w h i c h can operate w i t h o u t submersion. If tu b ehe ad operates satisfactorily in this position then c h e c k oil level by removing cover and oil filling p l u g — b e i n g careful that head is t i p p e d up so that oil filling nole is at highest level. When removing oil filling plug, finger should be held over expansion bag oil filling hole so that when v a c u u m is broken, that oil will not fill expansion bag compl etel y. Keep bag oil filling hole below level of oil filling plug or air will get into expansion bag. V 1.6 amps. 4.7 amps. 9.5 amps. page nine 63 72 80 CIRCUIT BREAKER OPENS REPEATEDLY 1. Check item 1 above (Low oil). 3. It may also be caused by a high voltage short in the x-ray head other than the x-ray units. 4. If the x-ray head were to lose an excessive amount of oil as a result of an oil leak, it w o u l d arc over, causing the circuit breaker to kick out. 5. The above conditions will be e v i d e n c e d by a tendency for the pointer on the milliameter to fly across the scale and go back t o " O " a s the circuit b r e a k e r opens. 6. It may indicate a short in the control or within the cable between the control and s h o c k p r o o f head. To eliminate the s o u r c e of trouble being in the head, d i s c o n n e c t X1 and X2 from the head. Tape the e n d s of the terminals and try to operate the unit again. If the circuit breaker does not kick out, then the trouble is in the x-ray head. C H E C K I N G TIMER Periodically c h e c k the timer for a c c u r a c y and also to be assured that the timer knob or pointer set screw has not loosened a n d is indicating the wrong time setting. HOW TO A D J U S T THE TIMER The timers used with these units contain means to compensate for contactor lag, and the adjustment is quite simple. Set the timer dial to 1/15 second and adjust t i m e r to obtain 4 dots with a spinning top. (See F i g . 2) This adjustment will not affect the longtime calibration by any significant amount. " L I T T L E GIANT 3 0 " a small unit doing a BIG J O B Universal X-Ray section four Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 s e c t i o n five MAINTENANCE FUSE L O C A T I O N S Cat. 3335 controls contain five fuses. Three fuses clip into a triple f u s e b l o c k mounted on the chassis. Use 3 a m p or 8 amp 250 volt 3AG fuses for replacement as indicated on the chassis beside the fuseblock. The remaining two fuses mount into fuseclips located on the rear of the Universal solid state timer circuit board and readily accessible from the rear. Use 8 / 1 0 amp. 250 volt 3AG fuses for replacement. C O L L I M A T O R 24 V O L T POWER SOURCE Terminals C1 and C2 in the control provide 24 volt (190 watt) ac source necessary to _ perate the light beam lamp in the collimator. MISCELLANEOUS C H E C K I N G TIMER Periodically c h e c k the timer for a c c u r a c y and also to be assured that the t i m e r knob or pointer set screw has not loosened and is indicating the w r o n g time setting. TIMER C A L I B R A T I O N The timer calibration can be c h e c k e d by any of the well known methods. Perhaps the simplest is to use a spinning t o p f o r short times and a cycle c o u n t e r f o r long times. Check the timer at 1 / 2 0 , 1 / 2 , 1 and 5 seconds. If exposure time deviates f r o m preset time by more than 1 cycle f o r 1 / 2 0 and / 2 second settings, or by more than 1 0 % for the 1 and 5 s e c o n d settings,, recalibrate timer. To recalibrate t i m e r see Fig. 2 under heading HOW TO A D J U S T THE TIMER. 1 SHORT IN mA. METER CIRCUIT . If mA. meter b o u n c e s around or does not read although there seems to be current flow t h r u the tube then c h e c k for a shorted mA. meter, short in wiring to the mA. meter or a short in the mA. to g r o u n d meter protection gap located at the transformer terminals. KILOVOLTMETER C A L I B R A T I O N This test requires a 150 volt ac voltmeter having an a c c u r a c y of 1% or better, and of known calibration. Connect the test voltmeter across the terminals of kilovoltmeter in the c o n t r o l . T u r n on the machine and vary the voltage by means of the kilovolt control selector s w i t c h . The voltage readings on the test meter should be c o m p a r e d with the a.c. volts on the top red scale of the control voltmeter. M A X I M U M DEVIATION Assuming that the x-ray e q u i p m e n t is properly installed and maintained a c c o r d i n g to the instructions s u p p l i e d , and a c c e p t e d p r o c e d u r e s are f o l l o w e d . The m a x i m u m deviations given here f o r individual control settings are b r o a d and do not reflect on the ability of the x-ray unit to conform to the R e p r o d u c i b i l i t y requirements listed under the Radiation Control A c t of 1968, 78.213-2b(1) Coefficient of v a r i a t i o n defined as, " F o r any specific c o m b i n a t i o n of selected t e c h n i q u e f a c t o r s , the estimated coefficient of variation of radiation exposures shall be no greater than 0.05. KILOVOLTS Kilovolts delivered to the x-ray t u b e will not deviate f r o m the indicated value by more than ± 2 0 % of the m a x i m u m kVp. 'LLIAMMETER C A L I B R A T I O N This test requires a dc milliammeter having a full scale reading approximately the same as the control milliammeter, with an a c c u r a c y o f 1 % o r better, and o f known calibration. Connect the test milliammeter directly in series with the control milliammeter by removing the lead f r o m one of the control milliammeter terminals and connecting that lead to the test milliammeter. Then connecting together the empty terminals of the two milliammeters. Read both meters at each of the milliammeter selector values and record the control milliammeter error. If the error e x c e e d s 5% of full scale, replace the milliammeter. QUICK ACTION METER A V A I L A B L E Universal has available a Cat. 3700 q u i c k action (approx. Va second) zero overshoot, test meter w h i c h can be purchased for servicemen to use in the field ( p r i c e $50 net). This will assure obtaining a more accurate reading with relatively short exposures, short enough not to overload the tube. EXPOSURE TIME Time of exposure will not deviate f r o m the value preset on the t i m e r scale by more than the f o l l o w i n g : 1/60 to 1/10 s e c o n d — 1 c y c l e 3 / 2 0 to 6 s e c o n d — 1 5 % MILLIAMPERES T h e milliamperes passing t h r u the X-Ray tube will not deviate f r o m the indicated value by more than 2 0 % of full scale. MISCELLANEOUS NOTE: T h i s section contains special a d d i t i o n s and c o r r e c t i o n s to assembly, o p e r a t i o n and maintenance instructions and s h o u l d be c o n s u l t e d before proceeding with assembly. This s e c t i o n five provides a means for e x p a n d i n g the usefulness of this manual by insertion of newly developed data gained f r o m field e x p e r i e n c e and for recording data d e v e l o p e d by the assembler or user a p p l i c a b l e to local conditions. C H E C K I N G INVERSE SUPPRESSOR IN 30 MA. C O N T R O L When a t u b e h e a d fails in use the inverse s u p p r e s s o r may be subjected to an unusually high c u r r e n t d r a w and thus short out. T h e c o n t r o l w o u l d still function even with a s h o r t e d d i o d e and w o u l d only manifest itself by a h i g h e r than normal high tension p r i m a r y c u r r e n t draw. It is s u g g e s t e d that the 30 mA. control be c h e c k e d w i t h the Cat. #3205 t u b e h e a d to establish that the inverse suppressor d i o d e is f u n c t i o n i n g and not shorted out. Use a clip-on ammeter in the X1 c i r c u i t (the t e r m i n a l s under the t u b e h e a d c o v e r are m a r k e d ) . Set the control for 5mA. C h e c k the c l i p on a m m e t e r readings at M, Q and V setting of kilovoltage. Y o u should get a p p r o x i m a t e l y 3.5 to 4 a m p e r e s . It should be a b o u t the same at all the kV. settings. A h i g h e r reading, a p p r o x i m a t e l y double w o u l d indicate a possibility of a shorted d i o d e in the control c o n t a c t o r circuit (inverse s u p p r e s s o r ) . See wiring d i a g r a m . Replace the diode if s h o r t e d . The diode values are 400 volts, 25 amperes. In rare instances and because of i m p r o p e r field c h a n g e s made on the control or t u b e head it is possible for the diode to be in g o o d c o n d i t i o n and yet a higher than normal c u r r e n t exists in the X1 lead. This may be c a u s e d by i m p r o p e r phase relationship of the d i o d e and the x-ray t u b e h e a d . It can also be caused by shorted turns in the s e c o n d a r y or a high resistance short s o m e w h e r e in the hi-volt circuit. Factory repairs or replacement is necessary. For p r e c i s e evaluation of t u b e h e a d o n l y see p a g e 5. Chart on page 5 provides normal c u r r e n t values measured in the X1 lead (primary of H.T.) of the t u b e h e a d . See SECTION FIVE f o r more " T r o u b l e s h o o t i n g " information that may have been a d d e d later to this m a n u a l . p a g e ten Universal X-Ray " L I T T L E G I A N T 30" a small x-ray unit d o i n g a BIG JOB Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 s e c t i o n five MISCELLANEOUS CAUTION DURING C O L D WEATHER SHIPMENT OF X - R A Y HEADS Exposure to c o l d weather during shipment may cause oil in x-ray head to contract, creating a condition of " l o w in o i l , " w h i c h can cause arc over or tube puncture, and d a m a g e to x-ray head. Slowly shaking head b a c k a n d f o r t h and listening f o r a splashing of oil will determine if head is low in o i l . A splashing sound can also be caused by a tube b r o k e n in shipment. This can readily be d e t e r m i n e d by making an x-ray exposure at the lowest setting of Pkv. Low oil will cause arcing and resultant high current draw. Shipment and outside storage during extreme c o l d w e a t h e r can cause insulating oil in t u b e h e a d to contract beyond the volume of oil retained in the expansion b a g . REVISED PANEL LAYOUT INTRODUCED APRIL, 1979 Carton may act as an insulator—therefore remove f r o m c a r t o n and permit e x p o s e d t u b e h e a d to w a r m up to room t e m p e r a t u r e . This may require 24 hours or longer. C h e c k o u t at low values of kilovolts and gradually bring up to full value. To mark preset volts and knob position on chart for 15, 20, 25, and 30 mA. settings respectively, one of two methods can be used: Be sure t u b e p o r t is covered w i t h a d e q u a t e sheet lead or that c o l l i m a t o r o p e n i n g i s c l o s e d and o r covered w i t h sheet lead t o prevent e x p o s u r e t o radiation during testing. BASIS OF MEASUREMENTS OF KVP., AND TIME MA., The basis upon which the maximum deviation of kVp. is stated is a sphere gap test, or a calibrated high voltage bleeder unit such as the Machlett Dynalyzer. The measurement basis upon which the maximum deviations of mA. are stated is a D.C. milliameter calibrated against a laboratory standard meter, in series with the x-ray tube. The measurement basis upon which the maximum deviations of exposure time is stated are half cycle impulses of the 60 hertz power line measured at the primary winding imput to the high voltage transformer or in the secondary circuit by means of a high voltage bleeder unit and an oscilloscope. page eleven 1. Lay a strip of transparent tape over the chart area (tape such as scotch transparent), and indicate values with a pen. To prevent smudging, lay a second strip of tape over the first, or; 2. Using a permanent dry marker, (such as Carter's "Marks A Lot") write in the values on the chart. These values can be removed with alcohol without damaging the paint on the control panel. LINE VOLTS 97 104 110 117 123 130 205 211 217 224 230 237 243 250 CONNECT L1 to 1 1 1 1 1 1 0/S1 0/S1 0/S1 0/S1 0/S1 0/S1 0/S1 0/S1 CONNECT L2 to 9 10 11 12 13 14 7 8 9 10 11 12 13 14 30 MA. "LITTLE GIANT CONTROL CAT. NO. 3335 SCHEMATIC - SEE 3335 CONTROL WIRING DIAGRAM FOR DETAILS Universal X-Ray Division of Chemetron Corporation 4014 West Grand Avenue Chicago, Illinois 60651 U.S.A. Telephone: (312) 276-4487 Cable UNIVAPROCO #040976 CUT ALONG DOTTED LINE AND PUT UNDER PLASTIC ON FRONT OF #3335 CONTROL USE THIS CHART TO OBTAIN THE ACTUAL VALUES OF KILOVOLTAGE UNDER LOAD AT A PARTICULAR MILLIAMPERE (MA.) SETTING. N 0 P Q R S T U V W X No Load Primary (Ep)Volts 60 65 71 77 83 88 95 102 108 115 123 129 Load Pri. Volts 58 at 1 5 mA. kVp. (kilovolts) 40 63.5 69 75 81 87 93 99.5 106 112 118.5 124.5 44 50 55 60 64 69 74 80 85 92 Load Pri. Volts 20 mA. kVp. (kilovolts) Load Pri. Volts at 25 mA. kVp. (kilovolts) Load Pri. Volts at 30 mA. kVp. (kilovolts) 57 62.5 68.0 73.5 80 86 92 98 105 111 117 32 39 44 48, 53 59 64 71 76 80 85 62.5 68 73.5 79 85 91 98 104 28 34 38 44 50 54 59 65 71 76 56 62 67.5 73 78.5 84.5 90.5 96,5 103 109 24 30 35 40 45 50 55 60 65 72 KV KNOB SETTING at M 56.5 109.5 96 123 91 1.16 122 80 85 115.5 76 121.5 81 It is recommended that ONE value of mA. be selected and used and the time varied for variations in density requirements (mAs.). A FIXED value of kilovoltage can be selected to provide the desired contrast and exposure latitude. The milliamperes (mA.) setting is preestablished by recording the PRESET VOLTS. An approximation of mA. can be obtained by setting the mA. knob to a prerecorded position. An accurate presetting of mA. is made by setting the mA. knob to the recorded PRESET VOLT reading on the voltmeter. See manual for details. Preset Volts Knob Position Preset Volts 15 mA. 25 mA. 20 mA. 30 mA. Universal X-Ray Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 Knob Position FOR REVISED PANEL LAYOUT INTRODUCED APRIL, 1979 SEE PAGE 11. Universal X-Ray Division of Chemetron Corporation 4 0 1 4 West Grand Avenue Chicago, Illinois 60651 Telephone: (312) 276-4487 MAINTENANCE S C H E D U L E FOR R A D I A T I O N S A F E T Y OF X - R A Y APPARATUS In order to assure continued compliance to the Federal Performance Standards, maintenance inspections and tests should be carried out by qualified personnel on original installation and at intervals of 6 months or less. Specific instructions for performing these maintenance activities are provided by the individual manufacturers. This schedule relates- to conformance to the Federal Performance Standards and is not intended to assure general equipment performance which must be carried out independently. TUBE ASSEMBLIES: 1. 2. 3. Radiation Leakage. Look for obvious physical damage which would affect radiation shielding and proper beam limiting device function. Beam quality. Confirm that the minimum filtration required is in the useful beam. Check interlocks where applicable. Field limitation and alignment. Check indicators of beam axis and centering. 9. 10. FLUOROSCOPIC IMAGING ASSEMBLIES - NON IMAGE INTENSIFIED: 1. Radiation leakage. Look for obvious physical damage which would affect radiation shielding. 2. Primary protective barrier. Check that the entire useful beam is intercepted by the primary barrier at any S.I.D. and that the fluoroscopic tube shall not produce x-rays if the barrier is not in the intercepting position. Check to be sure lead glass has not been unknowingly replaced by ordinary plate glass. 3. GENERATORS and CONTROLS: 1. 2. Line voltage and voltage regulation. Measure line voltage and voltage regulation to confirm that both are within manufacturer's specifications and that supply line is connected to the proper line terminals, if applicable. Calibration. a) Perform the manufacturer's calibration procedure, including, but not necessarily limited to, testing and adjusting kV, mA, mAs, and timer factors to specifications in manufacturer's instructions. b) Confirm that exposure cannot be made if timer is not set on a specific time or if timer is set to "zero", if applicable. 3. Contactors and relays. Inspect electromechanical contactors and relays for pitting, poor contact, loose, or missing parts. Replace if necessary. 4. High tension cable and transformer bushings. Check for proper filament circuit contact to insure consistant mA. output. 5. Collimator filter interlock. Confirm, where applicable, that an exposure cannot be activated at 50 kV and above, if the minimum required filtration is not in place. Visual exposure indicator. Confirm that the means provided and specified by the manufacturer for indicating, visually, the occurrence of an x-ray exposure (mA meter, pilot light, etc.) is functioning during and only during an exposure. 6. 7. 8. Audible exposure indicator. Confirm that the audible indicator provided by the manufacturer to indicate the termination of an x-ray exposure is functioning in the manner specified by the manufacturer and only in the manner specified. Inspect and test dials and knobs. Inspect knobs on timers, kilovoltage, millampere selectors and any other adjusting knobs to be sure that the pointer is indicating to the proper value. X-ray exposure switch. Confirm that any switch provided for activating an x-ray exposure requires continuous pressure to maintain the exposure or that release of the switch terminates the exposure. Warnings and indicators legible. Inspect and confirm that all warning labels and embossed, painted, silk screened, or other wearable technique factors indicators have not been defaced or worn so as to be illegible. Check that the means provided to further limit field size function properly, and that a minimum field size of 5 x 5 cm. at maximum S.I.D. can be provided. 4. Fluoroscopic exposure rate. Measure fluoroscopic radiation dose rate and adjust, if necessary, to not more than 5 R/ minute or not more than 10 R/minute, depending on equipment. Check B.R.H. requirements. TABLES: 1. Aluminum equivalent. Inspect table tops and cradles for any physical damage, alterations or deviations from the certified model which might alter the attenuation characteristics. CRADLES: 1. Beam attenuation. Check that the cradle has not been modified or changed , that the aluminum equivalence exceeds 2.0 mm. CASSETTE HOLDERS: 1. Inspect the front cover, if provided, of the cassette holder for any physical damage or modifications which would alter the attenuation characteristics. 2. Interlocks. Test and confirm the proper operation of interlocks, if provided for the operation of positive beam limitation. Alignment indicators. Test and confirm the proper operation and accurate indications of means provided to accomplish alignment between the x-ray field and the image receptor. Alignment in Positive Beam Limitation (P.B.L.). If provided for operation with positive beam limitation, test and confirm proper°alignment between the x-ray field and the image receptor. 3. 4. (continued from preceding page) MAINTENANCE S C H E D U L E FOR R A D I A T I O N S A F E T Y O F X - R A Y APPARATUS BEAM LIMITING DEVICES: 1. Leakage Radiation. Inspect beam limiting device and its attachment to the x-ray tube housing for physical damage, loosening or wear which might affect leakage radiation. Verify that the combination of tube housing and beam limiting device is listed as compatible. 2. 3. 4. Beam quality. a) Verify that all filtration elements, incidental and added, as provided by the certified beam limiting device and tube housing, are present and show no evidence of physical damage or alteration which might alter attenuation. b) Verify operation of filter - kV interlock in systems with more than one thickness of filtration. c) Perform such maintenance as specified by manufacturer, e.g., tightening of hardware, lubrication, etc. Variable field. Verify functioning of stepless adjustment of x-ray field size. Verify that minimum field size of 5 x 5 cm. can be achieved at 100 cm, Visual definition. a) Verify that misalignment between the visually defined field and the x-ray field does not exceed 2% of S.I.D. b) Verify average illumination exceeds 160 lux (15 foot candles). c) 5. 6. • 7. 8. Intraoral image receptors. a) Verify presence and integrity of source to skin distance limiting devices specified for the certified component. b) Verify maximum field size at minimum SSD is 7 cm. or 6 cm., respectively, if minimum SSD is greater than or less than 18 cm. Single image receptor size. Verify presence, integrity and functioning of means provided to limit x-ray filed size to not greater than the image receptor, and to align the field and receptor centers to within 2% of S.I.D. 9. Other x-ray systems. Verify presence, integrity and functioning of means specified and provided for the certified component to limit the x-ray field such that each dimension does not exceed image receptor size by more than 2% of S.I.D. and to align the centers of the x-ray field and receptor to within 2% of S.I. Verify presence and visibility of, markings identifying image receptor size and S.I.D. 10. Field limitation and alignment. a) Verify that the total misalignment of x-ray field with the respective edges of the selected portion of the image receptor does not exceed 3% of S.I.D. for length or width, nor 4% of magnitudes for the two directions. b) Verify that edge contrast ratio of light field exceeds 4 and 3 respectively for stationary and mobile radiographic equipment. d) Perform manufacturers routine maintenance. Field indication and alignment. a ) " Verify the proper functioning of the means for alignment of the center of the x-ray field with the center of the image receptor. b) Verify that the numerical indications of field size result in x-ray field dimensions in the plane of the image receptor are within 2% of S.I.D. of the dimensions of the•image receptor. c) Perform manufacturers routine maintenance. Positive beam limitation. a) Verify automatic adjustment of x-ray field size to image receptor size within 5 seconds of insertion of image receptor, or inhibition of exposure until field congruency is obtained, b) Verify that the x-ray size conforms to that of the image receptor within 3% of S.I.D. per axis and 4% of S.I.D. total. c) Varify operation of optional field size reduction and that field can be reduced to 5 x 5 cm. or less at 100 cm. d) Verify that return to positive beam limitation occurs upon a change in image receptor. e) Verify that the bypass mode, where provided, functions when not using the cassette tray or permanently mounted vertical cassette holder, and when either beam axis or table angulation if not within 10° of the horizontal or vertical during any part of the exposure. Verify automatic return to positive beam limitation when more of the above are. applicable. f) Verify operation of override key, where provided. c) d) 11. Verify that x-ray field size can be adjusted smaller than the selected portion of the film, and that at maximum S.I.D. a field size of less than 5 x 5 cm. can be provided. Verify that the centers of the x-ray and selected film portion coincide within 2% of S.I.D. Perform manufacturers routine maintenance. Source skin distance (mobile units). Verify the presence and integrity of the means specified and provided for the certified component to limit source skin distance to no less than 30 cm. Check that the certification label and component identification label are affixed to all certified components. Consult DHEW Publication No. (FDA) 75-8003 dated July, 1974 for specific compliance requirements. Also consult B.R-.H. Routine Compliance Testing for Diagnostic X-Ray Systems or Components of Diagnostic X-Ray Systems, to which 21 CFR Subchapter J is applicable, DHEW Publication (FDA) 75-8012 dated September, 1974 for details on test procedures and equipment specifications,and details. After initial installation, performance of these periodic activities is the responsibility of the owner/user, as they are not provided at no cost under the manufacturer's warranty.