Uploaded by CHAFFEEJAXON

11 Noninvasive Monitoring

advertisement
Non-Invasive
Monitoring
CRITICAL CARE TECHNIQUES
Components of NonInvasive Monitoring

Pulse-oximetry

Capnography

Transcutaneous Monitoring (a.k.a. T-Com)
Pulse Oximetry

Pulse oximetry provides an estimate of SaO2
by utilizing selected wavelengths of light to
non invasively determine the saturation of
pulsed blood (SpO2)

Uses two LEDS to transmit light through
capillary bed to photodiode detector at
different wavelengths.

Ex.

One diode (Red) – emits light at 660nm

Second diode (infared)– emits light at 940nm
(Lights flashing several hundred times per
second!
3
Principle of absorption
Pulse Oximetry

Indications
Monitor - the adequacy of arterial
oxyhemoglobin saturation
 Quantify response - of arterial oxyhemoglobin
saturation to therapeutic intervention or to
diagnostic procedures, such as bronchoscopy
 To comply - with mandated regulations or
recommendations by authoritative groups


Contraindications

The ongoing need for actual measurements of
pH, PaCO2, total hemoglobin, and abnormal
hemoglobins may be a relative contraindication
to pulse oximetry
5
Pulse Oximetry

False high or low readings can be
caused by several factors.

Factors affecting SpO2 accuracy include

Motion

Abnormal hemoglobin such as
CARBOXYHEMOGLOBIN!!! POX will read high!

Intravascular dyes

Skin pigmentation

Nail polish

Low perfusion states – such as cold hands, low BP, &
hypovolemia.
6
Pulse Oximetry (Cont.)
7
Pulse Oximetry

Remember
 POX
does not measure CaO2 or PCO2;
patients suspected of having O2
transport issues or hypoventilation
should have an ABG
8
Capnography
Defined as the measurement of C02 in an
expired gas.


Two main types of Capnography
 Qualitative
 Quantitative
Qualitative Devices

Also referred to as “chemical” capnometry

Color change on litmus denotes at least a 5%
change in C02 levels.


Used in RSI
Not for trending
Quantitative
Capnography

Take sample of exhaled gas and provides a
number display and graphic display of results

Two types exist

Side stream

Mainstream
Mainstream vs. Sidestream

Sidestream


Sample is collected via a small port that feeds to
outside sensor (monitor).
Mainstream

Sample chamber attaches directly between ETT
and Y adapter, so there is a continuous reading of
PetC02
Mainstream vs. Sidestream
13
Mainstream
Sidestream
Nasal Cannula w/EtCO2
Continuous Waveform
Capnography

Can be breath to breath

This is the gold standard for endotracheal intubation


If a square is seen, the ETT cannot be anywhere other
than in the trachea. Auscultate bilaterally.
Can also be viewed as a trend

https://www.youtube.com/watch?v=_KKliTT3DSo
Normal SBc02 waveform
Normal value

End tidal C02 (PetC02)
◦ Normal = 35-40 torr
◦

a-e diff. (PaCO2-PetCO2)
Can be used to estimate dead space ventilation.
◦ Normal a-e diff. = 2-5 mmHg
◦ The a-e difference can never be 0

◦
https://www.youtube.com/watch?v=mrQ8pghnC94
Increased Vd

More wasted ventilation = more deadspace


results in a widened (a-e diff)
Example: PaC02 50 and PetC02 20

An a-e diff of 30 is huge!

https://www.youtube.com/watch?v=sB3J3DhhUT0
Widening a-e diff.

Example:


PaC02 is 45 and PetC02 is 30
difference is 15 torr (widened or increased)
Examples of when this may happen

COPD who is air-trapping

Pulmonary emboli

Over-distended alveoli or excessive PEEP or pulmonary HTN

LHF blood backs up to pulmonary capillary beds and less C02
is excreted
T-com Monitor
https://www.youtube.com/watch?v=7d7XNtQNllA
Transcutaneous
Monitoring
Provides continuous, noninvasive estimates of
PO2 and PCO2 using skin sensor
 More commonly used in neonates
 Sensor warms underlying skin to increase
arterial blood flow
 Two most important factors influencing
accuracy of transcutaneous measurements:
age and perfusion status



Low perfusion and increasing age reduce
agreement between PtcO2 and PaO2
Agreement between PtCO2 and PaCO2 is
better because CO2 is more diffusible through
skin
23
Transcutaneous
Monitoring

Most common sites for electrode placement for
infants and children are abdomen, chest, and
lower back

Should compare monitor readings with those
obtained with concurrent ABG

Validation with ABG should be repeated any time
patient’s status undergoes major change
24
Transcutaneous
Monitoring


Indications

The need to monitor continuously the adequacy of
arterial oxygenation or ventilation

The need to quantify the real-time responses to
diagnostic and therapeutic interventions, as
evidenced by PtCO2 or PtCO2 values
Contraindications

There are no absolute contraindications. In patients
with poor skin integrity or adhesive allergy,
alternative devices should be considered
25
Transcutaneous
Monitoring

Hazards and possible complications


False-negative or false-positive results may lead to
inappropriate treatment.
Tissue injury

Erythema (redness)

Blisters

Burns
26
Download