EN INSTRUCTION MANUAL FUJI DRI-CHEM CLINICAL CHEMISTRY ANALYZER FUJI DRI-CHEM NX600 FUJI DRI-CHEM NX600i This Manual describes details on how to operate the FUJI DRI-CHEM NX600/NX600i and cautions to be observed when operating it. Please read this manual thoroughly before actually operating the FUJI DRI-CHEM NX600/ NX600i. After reading this manual, store it nearby the FUJI DRI-CHEM NX600/NX600i so that you can see it whenever necessary. Safe Usage and Handling Precautions 1 Part Names and Consumables 2 Measurement 3 Applicable Usage Methods 4 Loading Consumables 5 Maintenance 6 Quality Control 7 Troubleshooting 8 Function Menu 9 Other Functions 10 USM 11 Specifications 12 Glossary 13 2nd Edition 897N203029A January 2022 Introduction Thank you for purchasing the FUJI DRI-CHEM NX600 or FUJI DRI-CHEM NX600i. Accompanying documents were originally drafted in the English language. Installation may only be conducted by authorized service personnel. <Intended Purpose> The FUJI DRI-CHEM NX600 and FUJI DRI-CHEM NX600i (henceforth, the FDC NX600) is an in vitro diagnostic medical device to automatically quantitate the concentration or the activity of the components in blood or urine by colorimetric end-point, rate and ISE tests, using the FUJI DRI-CHEM SLIDEs. Do not use the equipment for other purposes. This equipment must be operated by Healthcare professionals who are familiar with its intended purpose and correct operation. And, the equipment can be used in a near patient setting by Healthcare professional. CAUTIONS 1. No part or all of this manual may be reproduced in any form without prior permission. 2. The information contained in this manual may be subject to change without prior notice. 3. FUJIFILM shall not be liable for malfunctions and damage resulting from installation, relocation, remodeling, maintenance, and repair done by other than dealers specified by FUJIFILM. 4. FUJIFILM shall not be liable for malfunctions and damage of FUJIFILM products due to the use of products of other manufacturers not supplied by FUJIFILM. 5. FUJIFILM shall not be liable for malfunctions and damage resulting from remodeling, maintenance, and repair using repair parts other than those specified by FUJIFILM. 6. FUJIFILM shall not be liable for malfunctions and damage resulting from negligence of precautions and operating methods contained in this manual. 7. FUJIFILM shall not be liable for malfunctions and damage resulting from use under environment conditions outside the range of conditions required for proper use of this product, such as power supply, installation environment, etc. contained in this manual. 8. FUJIFILM shall not be liable for malfunctions and damage resulting from natural disasters such as fires, earthquakes, floods, lightning, etc. Regulation (EU) 2017/746 requires that any serious incident that has occurred in relation to the device should be reported to the manufacturer or European Authorized Representative, and the competent authority of your country. In case you become aware of any serious incident, please report immediately to your local or our official dealer and the competent authority of your country. Trademark “DRI-CHEM” is a trademark or a registered trademark of FUJIFILM Corporation. Other holder’s trademarks “QR Code” is registered trademark of DENSO WAVE INCORPORATED. * This product adopts the font of ARPHIC TECHNOLOGY CO., LTD. Copyright © 2020-2022 FUJIFILM Corporation. All rights reserved. 897N203029A FDC NX600 Instruction Manual 2022.01 (i) Introduction The FDC NX600 is a diagnostic medical device, and this manual is available with all the series (FUJI DRI-CHEM NX600 and FUJI DRI-CHEM NX600i; the former includes the PF (Plasma Filter) unit, and the latter does not. So for the latter, sections that describe the unit can be skipped). Name PF Unit FUJI DRI-CHEM NX600 4 FUJI DRI-CHEM NX600i - NOTE: 4 : The PF Unit is mounted on the device. - : The PF Unit is not mounted on the device. (ii) 897N203029A FDC NX600 Instruction Manual 2022.01 Table of Contents 1 Safe Usage and Handling Precautions 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 1.10 1.11 1.12 1.13 1.14 1.15 1.16 1.17 1.18 1.19 1.20 1.21 1.22 2 Part Names and Consumables 2.1 2.2 2.3 2.4 2.5 3 Precautions (marks) Used in This Manual……………………………………………… 1-1 Operation Precautions……………………………………………………………………… 1-1 Biohazards and Disposal…………………………………………………………………… 1-2 Warnings Regarding Explosives…………………………………………………………… 1-2 Warnings Regarding Electric Shock……………………………………………………… 1-3 Electromagnetic Compatibility (EMC)…………………………………………………… 1-3 Caution Regarding Moving Parts………………………………………………………… 1-4 Installation Locations……………………………………………………………………… 1-4 Action for Malfunction……………………………………………………………………… 1-5 Measurement Results……………………………………………………………………… 1-6 FUJI DRI-CHEM Slides…………………………………………………………………… 1-6 CRP Measurement………………………………………………………………………… 1-7 QC Card System…………………………………………………………………………… 1-7 Handling Samples…………………………………………………………………………… 1-7 Sampler Cover……………………………………………………………………………… 1-8 Diluents and Reference Fluid……………………………………………………………… 1-8 FUJIFILM PF Plasma Filters……………………………………………………………… 1-8 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and FUJI DRI-CHEM MIXING CUPS……………………………………………………… 1-8 Light Source Lamp………………………………………………………………………… 1-9 Recording Paper…………………………………………………………………………… 1-9 Labels………………………………………………………………………………………… 1-10 Symbols……………………………………………………………………………………… 1-11 Part Names………………………………………………………………………………… 2-1 Touch Panel Screen Names and Functions……………………………………………… 2-5 2.2.1 Measurement Preparation Screen (Top Screen)……………………………… 2-5 2.2.2 Help Screen………………………………………………………………………… 2-6 Names and Functions of Software Keyboard Screens………………………………… 2-8 Bundled Items……………………………………………………………………………… 2-10 Consumables and Optional Items………………………………………………………… 2-11 2.5.1 Consumables……………………………………………………………………… 2-11 2.5.2 Optional Items……………………………………………………………………… 2-12 Measurement 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 Operation Overview………………………………………………………………………… 3-1 Measurement Preparation………………………………………………………………… 3-3 Basic Measurements……………………………………………………………………… 3-7 3.3.1 Lamp Off and Measurement-start Scheduling………………………………… 3-14 Measurement Results Screen……………………………………………………………… 3-15 ISE Tests……………………………………………………………………………………… 3-18 Dilution Tests………………………………………………………………………………… 3-21 CRP Tests…………………………………………………………………………………… 3-25 Shutting-down the FDC NX600…………………………………………………………… 3-27 897N203029A FDC NX600 Instruction Manual 2022.01 i Table of Contents 3.9 Sample Tubes for the FDC NX600………………………………………………………… 3-30 3.9.1 Blood Collection Tubes…………………………………………………………… 3-30 3.9.2 FUJI Sample Tubes……………………………………………………………… 3-32 3.9.3 Assembling the Sample Rack…………………………………………………… 3-33 3.10 QC Card System…………………………………………………………………………… 3-37 4 Applicable Usage Methods 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 5 Loading Consumables 5.1 5.2 5.3 5.4 5.5 6 Loading Consumables……………………………………………………………………… Setting Tips…………………………………………………………………………………… Loading Mixing Cups……………………………………………………………………… Loading Diluent……………………………………………………………………………… Loading Reference Fluid…………………………………………………………………… 5-1 5-2 5-5 5-7 5-9 Maintenance 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 ii CRP Calibration……………………………………………………………………………… 4-1 4.1.1 Calibration Measurement………………………………………………………… 4-1 4.1.2 Configuring Calibrator Concentration…………………………………………… 4-10 Manual Spotting (Colorimetric Slide)……………………………………………………… 4-13 4.2.1 Measuring by Manual Spotting (Colorimetric Slide)…………………………… 4-13 4.2.2 Using a Pipette…………………………………………………………………… 4-17 Tests Using a Plasma Filter (PF)………………………………………………………… 4-18 4.3.1 Measurement……………………………………………………………………… 4-20 4.3.2 Blood Collection Tubes for Tests Using a PF…………………………………… 4-23 Retesting……………………………………………………………………………………… 4-24 Slide Loading Order When Measuring Different Items Simultaneously……………… 4-27 Automatic Acquisition of Sample Information (Work Lists)……………………………… 4-28 Editing Sample Numbers and Sample IDs……………………………………………… 4-30 Using the Sample Barcode Reader (Sold Separately)………………………………… 4-31 Periodic Maintenance Items……………………………………………………………… 6-2 6.1.1 Items for Daily and Periodic Inspection by the User…………………………… 6-2 6.1.2 Items for Maintenance and Inspection by Repair Personnel………………… 6-3 Cleaning the Air Filters……………………………………………………………………… 6-4 Inspecting and Cleaning the Transfer Bar, Incubator, Sample Application Unit, and ISE Unit………………………………………………………………………………… 6-6 6.3.1 Reference Plate Level Check…………………………………………………… 6-7 6.3.2 Cleaning the Transfer Bar, Incubator, Sample Application Unit, and ISE Unit………………………………………………………………………… 6-9 6.3.3 Reassembling Parts and Ending Cleaning……………………………………… 6-17 Replacing Recording Paper………………………………………………………………… 6-18 Replacing and Cleaning the Light Source Lamp………………………………………… 6-19 Maintenance of Sampler O-ring…………………………………………………………… 6-22 6.6.1 Cleaning the Sampler O-ring……………………………………………………… 6-22 6.6.2 Inspecting the Sampler O-ring…………………………………………………… 6-24 6.6.3 Exchanging the Sampler O-ring………………………………………………… 6-27 Cleaning the Slide Reading Section……………………………………………………… 6-30 Cleaning and Replacing the Reference Fluid Cap Packing…………………………… 6-31 PF Unit Maintenance……………………………………………………………………… 6-33 6.9.1 Cleaning the PF Unit (Sensor, Reference Plate, and Suction Pad)………… 6-33 6.9.2 Inspecting the PF Suction Pad…………………………………………………… 6-36 6.9.3 Replacing the PF Suction Pad…………………………………………………… 6-38 6.9.4 Replacing the PF Pump Tube…………………………………………………… 6-41 897N203029A FDC NX600 Instruction Manual 2022.01 Table of Contents 7 Quality Control 7.1 7.2 7.3 7.4 7.5 7.6 8 7-1 7-1 7-1 7-1 7-1 7-2 Troubleshooting 8.1 8.2 8.3 9 Control Fluids………………………………………………………………………………… Measurement with Control Fluids………………………………………………………… Monitoring Results………………………………………………………………………… Control Mode………………………………………………………………………………… Troubleshooting Quality Control Results………………………………………………… Control Chart………………………………………………………………………………… Error Display………………………………………………………………………………… 8-1 8.1.1 Error Log…………………………………………………………………………… 8-2 8.1.2 Error Code and Reference Section Table……………………………………… 8-4 8.1.3 Printout Reference Table………………………………………………………… 8-8 Troubleshooting……………………………………………………………………………… 8-10 8.2.1 Startup Errors……………………………………………………………………… 8-10 8.2.2 Printer or LCD Trouble…………………………………………………………… 8-10 8.2.3 Slide Reading Errors……………………………………………………………… 8-11 8.2.4 Sampler Errors……………………………………………………………………… 8-11 8.2.5 Photometric System Errors……………………………………………………… 8-19 8.2.6 Transfer Errors……………………………………………………………………… 8-21 8.2.7 Temperature Controller Errors…………………………………………………… 8-22 8.2.8 Circuit Board Signal Errors……………………………………………………… 8-24 8.2.9 QC Card, PF Card, and DI Card Errors………………………………………… 8-29 8.2.10 Slide Loading Errors……………………………………………………………… 8-31 8.2.11 ISE Measurement Errors………………………………………………………… 8-33 8.2.12 PF Errors…………………………………………………………………………… 8-36 8.2.13 Errors Related to Data Communication, Sample Barcode Reader…………… 8-37 8.2.14 Calibration Errors………………………………………………………………… 8-39 8.2.15 Other Errors………………………………………………………………………… 8-42 Slide Code Table…………………………………………………………………………… 8-44 Function Menu 9.1 9.2 9.3 Control………………………………………………………………………………………… 9-3 Normal Mode………………………………………………………………………………… 9-7 9.2.1 Date & Time Settings……………………………………………………………… 9-8 9.2.2 Brightness • Volume……………………………………………………………… 9-8 9.2.3 Lamp Management………………………………………………………………… 9-9 9.2.4 Lot Information……………………………………………………………………… 9-9 Mode Function for Administrator…………………………………………………………… 9-10 9.3.1 Language Setting………………………………………………………………… 9-11 9.3.2 Host Connection Setting………………………………………………………… 9-11 9.3.3 Calculated Parameter……………………………………………………………… 9-14 9.3.4 Unit Conversion…………………………………………………………………… 9-16 9.3.5 Switch Measurement Range Display…………………………………………… 9-17 9.3.6 Reference Interval Range Setting……………………………………………… 9-18 9.3.7 Sample No. /ID Settings…………………………………………………………… 9-20 9.3.8 Work List Selection Setting……………………………………………………… 9-20 9.3.9 Correlation Coefficients…………………………………………………………… 9-21 9.3.10 Lot Compensation Coefficients…………………………………………………… 9-25 9.3.11 Standard Curve Coefficients……………………………………………………… 9-27 9.3.12 Lot Qualification…………………………………………………………………… 9-30 9.3.13 PF Calibration Coefficients……………………………………………………… 9-31 9.3.14 Dilution Setting……………………………………………………………………… 9-33 9.3.15 Tube Setting for Diluent & Reference Fluid……………………………………… 9-35 9.3.16 Print Setting for Measured Result………………………………………………… 9-35 897N203029A FDC NX600 Instruction Manual 2022.01 iii Table of Contents 9.3.17 9.3.18 9.3.19 9.3.20 9.3.21 9.3.22 Control Fluid Setting……………………………………………………………… 9-36 Control Measurement Setting…………………………………………………… 9-37 Operator ID Setting………………………………………………………………… 9-37 Test Item Name Setting…………………………………………………………… 9-38 Default Sample Type Setting……………………………………………………… 9-39 Password Setting………………………………………………………………… 9-39 10 Other Functions 10.1 Data Communication……………………………………………………………………… 10-1 11 USM (User Safety Mode) 11.1 Operator ID Input When Measurement Start and In Mode Function………………… 11-1 11.2 Difference of Operable Functions among Operator ID Level (OPE,ADM,EMG)……… 11-2 11.3 USM Menu…………………………………………………………………………………… 11-5 11.3.1 Operator ID -Import………………………………………………………………… 11-5 11.3.2 Operator ID -Export………………………………………………………………… 11-7 11.3.3 Editing Registered Operator ID…………………………………………………… 11-9 11.3.4 Editing Emergency ID…………………………………………………………… 11-11 12 Specifications 13 Glossary iv 897N203029A FDC NX600 Instruction Manual 2022.01 1 Safe Usage and Handling Precautions This section contains the safety precautions that must be followed for safe operation of the FUJI DRI-CHEM NX600 (For the remainder of the manual, FUJI DRI-CHEM NX600 is shortened to “FDC NX600”). Before using this analyzer, please read this chapter carefully and follow the precautions given, so that you can operate it correctly. 1.1 1 Precautions (marks) Used in This Manual The following section lists the meaning of precautions (marks) used in this manual. WARNING: “Warning” indicates hazardous situations that may lead to serious injury, death, or the transmission of infectious agents if the precaution is not followed. CAUTION: “Caution” indicates hazardous situations that may lead to minor injury, moderate injury, or physical damage if the precaution is not followed. “Important” indicates improper handling that could have an adverse effect on IMPORTANT: the accuracy of the measurement values if the precaution is not followed. NOTE 1.2 “Note” indicates procedures requiring special attention, instructions that must be followed, supplementary explanations, etc. Operation Precautions CAUTION: Before using this equipment, please read this Instruction Manual carefully so that you can operate the equipment correctly. CAUTION: Whenever you operate this equipment, be sure to observe the precautions described in this manual. Failure to do so may subject you to injuries, cause property damage, or produce incorrect measurement results. CAUTION: The operators must be sufficiently trained for the usage of sampler cover lock key and observe its usage method and precautions. Store the lock key carefully so that it is not lost. (Refer to Section 1.15) CAUTION: Do not modify the FDC NX600. Otherwise, the safety will not be guaranteed. CAUTION: The effective period of service (service life) of this analyzer is 6 years after delivery (as long as usage precautions are observed and regular maintenance is performed). If you will use the analyzer after the service life expires, consult with the manufacturer/seller beforehand. The use of medical equipment that has exceeded its service life may increase the chance of safety and performance problems. IMPORTANT: Use consumables on or before their listed expiration date. 897N203029A FDC NX600 Instruction Manual 2022.01 1-1 1 Safe Usage and Handling Precautions 1.3 Biohazards and Disposal WARNING: Items such as measured slides, tips, mixing cups, and sample tubes coated with samples such as blood and urine, and cotton swabs and cloths used to clean the equipment are infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. Furthermore, when entrusting disposal to another party, request processing from a contractor with a license for special management of industrial waste processing work. 1 WARNING: Items that may be contaminated with blood or urine are infectious industrial waste. When discarding the main unit, follow applicable regulations in your country and dispose of it appropriately. WARNING: Process waste correctly, as stipulated by local law or any regulations that apply. When discarding the FDC NX600 that incorporates the lithium battery, be sure to contact a licensed waste disposal contractor because it cannot be disposed of as a general waste. Not doing so may cause environmental pollution. WARNING: Before discarding the main unit, you must delete the measurement results (personal information) stored on the main unit. Contact your dealer before discarding. WARNING: When handling samples (blood or urine) and performing maintenance (cleaning the analyzer), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Do not touch used slides and tips with bare hands as this may cause contamination. If you accidentally touch a contaminated consumable, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: When samples come in contact with the analyzer, immediately clean and disinfect it. 1.4 Warnings Regarding Explosives WARNING: As this equipment is not explosion-proof, be sure not to use flammable and explosive gas around the equipment. 1-2 897N203029A FDC NX600 Instruction Manual 2022.01 1 Safe Usage and Handling Precautions 1.5 Warnings Regarding Electric Shock WARNING: The voltage supplied to the equipment is AC 100 - 240 V. To avoid electrical shock, observe the following precautions. • Avoid installation locations in which water may splash on the equipment. • Make sure that the equipment is properly grounded to a protective earth lead for indoor wiring. • Make sure that all cables have been properly connected. WARNING: Plug the power cable of the equipment into an outlet with a grounding receptacle. Otherwise, there is risk of electrical shock. WARNING: When plugging the power cable in or removing it from an outlet, be sure to hold onto the plug body, not just the cable. Otherwise, the power cable may be damaged and cause electric shock or fire. WARNING: Do not remove covers or other parts that are secured with screws. Doing so may result in electrical shock from exposure to hazardous voltage or injury from moving parts. 1.6 Electromagnetic Compatibility (EMC) This equipment conforms to the following emission and immunity requirements: - IEC 61326-2-6:2012 (Class B) - EN 61326-2-6:2013 (Class B) - IEC 61326-2-6:2005 (Class B) - EN 61326-2-6:2006 (Class B) • The electromagnetic environment should be evaluated prior to operation of the equipment. • Do not use this equipment in close proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these may interfere with the proper operation. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures. • Reorient or relocate the receiving device. • Increase the separation between the equipment. • Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected. Consult the manufacturer or field service technician for help. CAUTION: Do not use other devices (such as mobile phone) which generate and can radiate radio frequency energy near the FDC NX600. Otherwise, physical damage or malfunction on the FDC NX600 may occur. NOTE It is the manufacturer's responsibility to provide equipment electromagnetic compatibility information to the customer or user. On the other hand, it is the user’s responsibility to ensure that a compatible electromagnetic environment for the equipment can be maintained in order that the equipment will perform as intended. 897N203029A FDC NX600 Instruction Manual 2022.01 1-3 1 1 Safe Usage and Handling Precautions 1.7 Caution Regarding Moving Parts WARNING: Do not place your fingers near moving parts (sampler and consumable disk) when operating the equipment, including during maintenance. Also be sure not to get your fingers, hair, clothing, or accessories caught in moving parts. 1 WARNING: During sample processing, ensure the sampler cover is closed and locked with the sampler cover lock key to prevent injuries and biohazards. If your finger touches a tip that has sample inside, the sample may scatter and cause infection. When the display reads [Ready to test], the sampler cover can be opened. 1.8 Installation Locations WARNING: Plug the power cable of the equipment into an outlet with a grounding receptacle. Otherwise, there is risk of electrical shock. CAUTION: Avoid the following installation sites. • Places where spills or water leakage may occur. • Places where the equipment is exposed to direct sunlight. • Places near sources of heat such as heaters. • Places where the temperature may drastically change. • Places where the equipment is subject to vibration or its support table is unstable. (1) Install the equipment in the following environmental conditions. Installation environment: Indoor use (do not expose to direct sunlight) Illumination: 6,000 cd/m2 (lux) or less(3,000 cd/m2 (lux) or less when using the sample barcode reader) Atmospheric pressure: 810 hPa or more (equivalent to altitude 2,000 m) Transient overvoltage category: II Pollution degree: 2 IP00 (No Protection) Ingress protection rating: Operating temperature: 15 to 32 °C Operating humidity: 30 to 80% RH (no vapor condensation) (2) Use the equipment under the following electrical specifications. single phase AC 100 - 240 V ±10% Power supply voltage: Frequency: 50 to 60 Hz Consumption current: 1.9 - 0.8 A Type of protection against electrical shock: CLASS 1 EQUIPMENT (3) Use an independent outlet separate from other equipment. (4) Make at least 10 cm of space behind and beside the equipment. (5) Unplug the equipment from the AC outlet if it will not be used for an extended period of time. 1-4 897N203029A FDC NX600 Instruction Manual 2022.01 1 Safe Usage and Handling Precautions CAUTION: Hold the handle of the device when lifting or moving it. 1 Handle CAUTION: • When moving the device to another installation location, remove all slides, samples, diluent, reference fluid, mixing cups, tips, and other consumables beforehand. • Do not set the device on its side or upside down while moving it. • Place the device on a flat and stable stand. • When the device is subject to vibrations, such as when being moved by vehicle, it is necessary to secure the moving parts within the device. Contact your dealer. • The weight of 28kg is supported by 4 portions of rubber feet. The load on each portion is approx. 7kg. Make sure that the load bearing of the table at the installation location is sufficient. Failure to observe these precautions may result in unexpected errors or equipment malfunctions. IMPORTANT: When wax was used to clean the floor of the installation location, make sure that wax will not adhere to slides and the device. If wax has adhered, the measurement results may be affected. 1.9 Action for Malfunction CAUTION: When the equipment generates unusual noise, smell, or smoke, immediately unplug the power cable, and contact your dealer. CAUTION: Do not place any objects near the wall outlet so that the power cable can be easily unplugged. CAUTION: In case of emergency, clearance at the back of the equipment is required so that the power cable can be removed from the inlet of the equipment. 897N203029A FDC NX600 Instruction Manual 2022.01 1-5 1 Safe Usage and Handling Precautions 1.10 Measurement Results IMPORTANT: Make a diagnosis in a comprehensive manner, considering other relevant test results or clinical symptoms. 1 IMPORTANT: When an equipment problem (error) appears before or during measurement, or a warning appears in the measurement results, the measurement results may not be accurate. Refer to Chapter 8, Troubleshooting, and then rerun the measurement. IMPORTANT: To maintain the accuracy of your test results, it is recommended to perform quality control on routine bases by use of control fluids. (Refer to Chapter 7.) 1.11 FUJI DRI-CHEM Slides NOTE For the remainder of the manual, FUJI DRI-CHEM slide is shortened to “slide”. CAUTION: Please confirm about the usable slide items in this device to your dealer for assistance. WARNING: Do not touch used slides with bare hands as this may cause infection. If you accidentally touch an item, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Used slides are infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. Furthermore, when entrusting disposal to another party, request processing from a contractor with a license for special management of industrial waste processing work. CAUTION: Please read the Instructions for Use of the slides carefully and follow the instructions. IMPORTANT: • The slides should be stored in a refrigerator still wrapped to avoid humidity, light, and heat. • Take out the required number of slides from the refrigerator, and allow them to reach room temperature for 5 minutes or more with the condition of the individual package before unwrapping and using them. Otherwise, the measurement results and the analyzer operation may be affected. • Complete the measurement within 30 minutes after unwrapping the individual package. • Do not touch the center part (both sides) of a colorimetric slide. • Do not touch the thread bridge part of an electrolyte slide. • Do not scratch the identification information on the back of a slide. • Do not reuse a slide. NOTE Slide packages and containers for liquids such as diluent and reference fluid may change without notice. 1-6 897N203029A FDC NX600 Instruction Manual 2022.01 1 Safe Usage and Handling Precautions 1.12 CRP Measurement WARNING: As calibrator CP is derived from blood, it should be handled in the same way as a patient sample. When handling calibrator CP, always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: CRP measurements require periodic calibration. For details, refer to the “Instructions for Use” of the slides and “Section 3.7 CRP Tests” in this document. 1.13 QC Card System (1) A QC card is packed together with CM (Colorimetric) slides in the same box. Before you start using a new lot of CM slides, read the QC card using the card reader. The analyzer can record up to 5 lots per test, so you can mix and use slides of different lots. Store the read QC card in the box it came in until the accompanying lot of slides has been used up. NOTE When QC information exceeds 5 lots, the oldest QC information is deleted. Slides without QC information cannot be measured. (2) A QC card contains the slide test name and the correction coefficient corresponding to the manufacturing number (lot). Always use slides from the same box. 1.14 Handling Samples WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: Sample preparations differ depending on the test. Read the “Instructions for Use” of the slides. IMPORTANT: When using plasma or serum, check that the samples have not undergone hemolysis, and that they do not contain any fibrin deposits. IMPORTANT: When using whole blood, test the sample as soon as possible after drawing it. In the case of whole blood, blood cell precipitation may affect the measurement results. 897N203029A FDC NX600 Instruction Manual 2022.01 1-7 1 1 Safe Usage and Handling Precautions 1.15 Sampler Cover WARNING: During sample processing, ensure the sampler cover is closed and locked with the sampler cover lock key to prevent injuries and biohazards. If your finger touches a tip that has sample inside, the sample may scatter and cause infection. When the display reads [Ready to test], the sampler cover can be opened. 1 1.16 Diluents and Reference Fluid CAUTION: To ensure safe handling of diluent and reference fluid, refer to the handling methods in the “Instructions for Use” of each fluid or follow the chemical handling rules of your facility (wear gloves, a lab coat, and safety goggles). If you accidentally touch an item, immediately rinse the contaminated body part thoroughly under running water. Seek medical assistance if necessary. IMPORTANT: For diluents used in measurements that require dilution, refer to the “Instructions for Use” of the slides. IMPORTANT: Use the reference fluid included with the ISE slides for ISE measurements. Refer to the “Instructions for Use” of the slides. IMPORTANT: Use only specified products. Otherwise, errors or malfunctions may occur. 1.17 FUJIFILM PF Plasma Filters NOTE FUJIFILM plasma filter PF is shortened to "PF" in this document. IMPORTANT: Before using a FUJIFILM PF, refer to “Instructions for Use” provided with the PF, paying special attention to the precautions concerning the samples measured and the blood collection tubes used. 1.18 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and FUJI DRI-CHEM MIXING CUPS NOTE In this manual, FUJI DRI-CHEM AUTO TIPS may be shortened to “tip”, FUJI Sample Tube to “tube”, and FUJI DRI-CHEM MIXING CUPS to “mixing cup”. FDC NX600 performs sample application automatically via the sampler. However, it is necessary to prepare dedicated FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and specified blood collection tubes and FUJI DRI-CHEM MIXING CUPS. The FUJI Sample Tubes include the following types. • FUJI HEPARIN TUBE 1.5 mL • FUJI HEPARIN TUBE 0.5 mL • FUJI PLAIN TUBE 1.5 mL • FUJI PLAIN TUBE 0.5 mL 1-8 897N203029A FDC NX600 Instruction Manual 2022.01 1 Safe Usage and Handling Precautions WARNING: Do not touch used tips, tubes, and mixing cups with your bare hands as this may cause infection. If you accidentally touch an item, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Used tips, tubes, and mixing cups are infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. Furthermore, when entrusting disposal to another party, request processing from a contractor with a license for special management of industrial waste processing work. IMPORTANT: Always use new FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, blood collection tubes, and FUJI DRI-CHEM MIXING CUPS for each sample. Do not reuse old ones. IMPORTANT: Use only specified products. Otherwise, errors or malfunctions may occur. IMPORTANT: When the power is shut off during test processing, such as by a power outage, replenish tips and replace mixing cups. The analyzer may lose the information of the usage positions of the consumables and use used mixing cups (holes). 1.19 Light Source Lamp CAUTION: The light source lamp gets very hot. Before replacing the lamp, turn the power off, wait at least 5 minutes, and then check that the lamp has cooled. (1) The light source lamp is a precision halogen lamp. Do not touch the glass surface of the lamp with your bare hands. (2) The lamp is a consumable with a lifetime. Always keep a spare lamp on hand in case the lamp burns out. 1.20 Recording Paper (1) Use specified recording paper for the FDC NX600. (2) If other recording paper is used, the printer head section may be damaged. 897N203029A FDC NX600 Instruction Manual 2022.01 1-9 1 1 Safe Usage and Handling Precautions 1.21 Labels The following safety labels are affixed to the FDC NX600. 1 Power button Sampler cover lock key caution label Biohazard label Refer to Section 1.15 Biohazard label High temperature caution label Lamp caution label PF tube caution label Biohazard label Refer to Section 6.5 Refer to Section 6.5 Moving parts caution label Refer to Section 6.9.4 Name-plate NOTE: The example above uses an illustration for the FDC NX600. For the FDC NX600i, the description of the plate partially differs. 1-10 Power switch | : Depress to power ON O : Depress to power OFF USB connector RS-232C connector LAN terminal 897N203029A FDC NX600 Instruction Manual 2022.01 1 Safe Usage and Handling Precautions 1.22 Symbols Sign Description 1 Warning, Caution, Important, consult documents Biological risks High temperature caution Power button Power switch LAN terminal RS-232C connector USB connector Fragile item, handle carefully Protect from rain This way up 2 Stacking limitation Temperature limitation Manufacturer SN Serial number 897N203029A FDC NX600 Instruction Manual 2022.01 1-11 1 Safe Usage and Handling Precautions Sign Description Date of manufacture 1 EC REP Authorized representative in the European Community In vitro diagnostic medical devices CE mark of conformity Importer Consult an electronic Instructions for Use (e-IFU) Unique device identifier Near-patient testing 1-12 897N203029A FDC NX600 Instruction Manual 2022.01 2 Part Names and Consumables 2.1 Part Names NOTE The following is an illustration of the FDC NX600 (full option model). FDC NX600i has slight difference. Printer 2 Sampler cover LCD (touch panel) Power button Sampler cover lock Front bottom cover Card reader Disposal box Sampler Incubator cable Plasma filter unit (PF unit) Incubator cover Consumable disk 897N203029A FDC NX600 Instruction Manual 2022.01 2-1 2 Part Names and Consumables Air filter (Louver) USB connector 2 (For sample barcode reader only) *Do not connect other USB devices. LAN terminal CAUTION: Do not remove the cover to look into the QR code reader’s illuminator. It could impair your eyes. RS-232C connector USB connector Power inlet Power switch Air filter (Louver) 2-2 897N203029A FDC NX600 Instruction Manual 2022.01 2 Part Names and Consumables 2 Disposal box Light source lamp Diluent holder Slide cartridge Reference fluid holder Sample rack RE Mixing cup Tip rack Sample application unit cover 897N203029A FDC NX600 Instruction Manual 2022.01 2-3 2 Part Names and Consumables < Details of PF unit > PF unit 2 PF sensor / suction pad PF reference plate PF sample rack RE PF pump 2-4 897N203029A FDC NX600 Instruction Manual 2022.01 2 Part Names and Consumables 2.2 Touch Panel Screen Names and Functions 2.2.1 Measurement Preparation Screen (Top Screen) If the power is switched on, the Measurement-preparation screen (top screen) is displayed after startup. 1 Help button 1 2 3 4 Display the help screen. è Refer to Section 2.2.2 2 2 Function menu button Display the function menu screen. è Refer to Chapter 9 3 Work list button 5 Obtain order information from the host PC via a DMS connection. è Refer to Section 4.6 4 Clear button 6 7 8 9 0 a CAUTION: The touch panel does not have multi-touch input function. Do not touch plural points on the panel at the same time. Clear the entered information. 5Plasma filter (PF) production number (lot) display Displays the lot number of the plasma filter (PF) in use. Make sure this number matches the production number (lot) of the plasma filter (PF) in use. 6 Mixing cups Displays the quantity of mixing cups. Tap to replace the mixing cups. è Refer to Section 5.3 7 Tips Tap to replenish tips. è Refer to Section 5.2 8 [START] button Start a measurement. 9O perator ID button (displayed according to settings) If tapped, you can enter the operator ID. è Refer to Section 9.3.19 0 Feed button Feed the recording paper a fixed length. [STOP] button (displayed during sample application operation) If the [STOP] button is tapped during sample application, tips are discarded, operation is stopped, and the device enters a paused state. a Measurement results button Display the measurement results. è Refer to Section 3.4 897N203029A FDC NX600 Instruction Manual 2022.01 2-5 2 Part Names and Consumables 2.2.2 Help Screen You can tap the button on the Measurement preparation screen to check the procedures about measurement, maintenance, consumable exchange, and other actions on the FDC NX600. 2 2-6 897N203029A FDC NX600 Instruction Manual 2022.01 2 Part Names and Consumables 1 1 button The Measurement-preparation screen appears. 2 button Return to the previous screen. 2 2 897N203029A FDC NX600 Instruction Manual 2022.01 2-7 2 Part Names and Consumables 2.3 Names and Functions of Software Keyboard Screens When data entry for operations and configurations is necessary, the software keyboard appears on the LCD automatically. Enter alphanumeric characters and symbols. 2 LCD (LCD touch panel) 1 2 3 1 Alphabetical characters, numbers, and symbols buttons Enter alphabetical characters, numbers, and symbols. 2 button Delete all entered characters, and then quit keyboard entry. 3 Move cursor buttons Move the position of the cursor left or right by one character. 4 [Clear] button Delete all entered characters. 7 6 4 5 [OK] button Confirm the entered characters. 5 6 button Delete the character to the left of the cursor. 7 [Space] button Enter a space of one single character width. 2-8 897N203029A FDC NX600 Instruction Manual 2022.01 2 Part Names and Consumables Change the entry screen by using the [ABC], [abc], [123], or [-#/], button. < When the [ABC] button is selected > < When the [abc] button is selected > < When the [123] button is selected > < When the [-#/] button is selected > 8 9 0 a 8 [ABC] button 0 [123] button 9 [abc] button a [- # /] button Display the entry screen for uppercase alphabetical characters and numbers. Display the entry screen for lowercase alphabetical characters and numbers. 897N203029A FDC NX600 Instruction Manual 2022.01 Display the entry screen for numbers, comma (,), and minus symbol (-). Display the entry screen for symbols and numbers. 2-9 2 2 Part Names and Consumables 2.4 2 Bundled Items Recording paper 2 rolls Slide cartridge 2 Slide weight 2 Light source lamp 1 O-ring 2 Tip rack 1 Sample rack assembly kit 1 set Sampler leak check tool 2 PF leak check tool (For FDC NX600) 1 FUJI DRI-CHEM AUTO TIPS 1 case FUJI PLAIN TUBE 0.5 (0.5 mL type) 1 pack FUJI DRI-CHEM MIXING CUPS 1 pack Manual pipette cover 1 European authorized representative label 1 Key (for sampler cover) 2 Instruction manual 1 Installation method 1 NOTE An AC power cable is not supplied as a bundled item. Please prepare an AC power cable conforming to the specifications below: <The requirements of the cable specifications for the FDC NX600> Power voltage: 100V-120V Requirements for the plug/connector: AC125V 10A Requirements for the cable: SVT 3/18AWG 60°C Power voltage: 200V-240V Requirements for the plug/connector: AC250V 10A Requirements for the cable: GTCE-3 1.0mm2 70°C NOTE The AC power cable is required to be compliance with any applicable regulations in your country. NOTE Do not change AC power cable to inappropriate one. NOTE Specifications and capabilities are subject to change without notice. 2-10 897N203029A FDC NX600 Instruction Manual 2022.01 2 Part Names and Consumables 2.5 Consumables and Optional Items To purchase the consumables or optional items listed below, please contact the dealer from whom you purchased the FDC NX600. 2.5.1 Consumables Name * FUJI DRI-CHEM AUTO TIPS FUJI Sample Tubes FUJI HEPARIN TUBE (1.5 mL) FUJI HEPARIN TUBE (0.5 mL) FUJI PLAIN TUBE (1.5 mL) * FUJI PLAIN TUBE (0.5 mL) * Recording paper * Light source lamp * O-ring Plasma Filter PF PF Suction Pad PF Tube * FUJI DRI-CHEM MIXING CUPS Reference fluid cap packing Package/unit 6 cases 2 500×1 box 500×1 box 500×1 box 500×1 box 6 rolls×1 box 1 4 50×1 bag 1 1 50×1 box 5×1 bag NOTE Parts names marked with “*” are the same parts packed with the FDC NX600. Chemicals FUJI DRI-CHEM CONTROL QP-L FUJI DRI-CHEM CONTROL QP-H FUJI DRI-CHEM CONTROL QN FUJI DRI-CHEM REFERENCE FLUID RE FUJI DRI-CHEM DILUENT DL (CRP) FUJI DRI-CHEM CALIBRATOR CP (CRP) NOTE Types of packages or containers for consumables are subject to change without notice. 897N203029A FDC NX600 Instruction Manual 2022.01 2-11 2 Part Names and Consumables 2.5.2 Optional Items Name * Sample rack assembly kit * Slide cartridge * Slide weight Sample barcode reader 2 Package/unit 1 set 1 1 1 CAUTION: The sample barcode reader specified for the FDC NX600 can be used. Do not connect a barcode reader other than specified for the FDC NX600. Otherwise, physical damage or danger of fire may result. NOTE Parts names marked with “*” are the same parts packed with the FDC NX600. NOTE Specifications and capabilities are subject to change without notice. 2-12 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 3.1 Operation Overview On the FDC NX600, first load the consumables and other items to prepare for measurement (Step 1), and then perform the actual measurement (Step 2). Step 1 - Measurement preparation Cleaning the disposal box 3 Refer to Section 3.2(1) Checking the recording paper Refer to Section 3.2(2) Starting the FDC NX600 Refer to Section 3.2(3) Checking the date and time Refer to Section 3.2(4) Loading consumables For dilution measurements For CRP measurement Diluent Diluent for CRP (Refer to Section 5.4) For ISE measurements Reference fluid (Refer to Section 5.5) 12 Tips (Refer to Section 5.2) 3-1 Refer to Chapter 5 For dilution measurements Mixing cups (Refer to Section 5.3) RE 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement Step 2 - Measurement Basic measurements 1 Refer to Section 3.3 2 Slide ISE test measurements Refer to Section 3.5 Dilution measurements Refer to Section 3.6 CRP measurements Refer to Section 3.7 Sample 3 RE Check the tips Check the sample type Plasma/Serum 3 START button Slide set order Refer to Section 4.5 To test CRP and CM (Colorimetric) slide To test CM and ISE slide To test CRP, CM, and ISE slide CRP ISE Slide CRP CM Slide CM Slide CM Slide ISE Slide To test CM slides that require dilution x5 Dil x2 Dil Checking the measurement results Refer to Section 3.3 (6) Ending measurement Refer to Section 3.3 (7) 897N203029A FDC NX600 Instruction Manual 2022.01 3-2 3 Measurement 3.2 Measurement Preparation (1) Cleaning the disposal box Empty the disposal box. IMPORTANT: If the disposal box is full, a slide transfer error or tip disposal error may occur, which may negatively affect the measurement values. 3 IMPORTANT: After emptying the disposal box, be sure to load the disposal box. If the analyzer is used without the box, the measurement results may be negatively affected. WARNING: Do not touch used slides and tips with your bare hands, as this may cause infection. If you accidentally touch an item, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Used slides and tips are infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. 3-3 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement (2) Checking the recording paper Press the area shown in the illustration to release the lock (1), and then lift up the cover (2). Cover 1 2 Press the concave section. 3 Check the remaining recording paper (FUJI thermal paper). If the recording paper is short for the measurement, replace the old roll with a new one. è Refer to Section 6.4 Recording paper NOTE Before closing the cover, pull out the edge of the recording paper. NOTE Close the cover securely until it clicks. Otherwise, the paper feed may fail and the test results will not be printed out. Recording paper Cutter opening 897N203029A FDC NX600 Instruction Manual 2022.01 3-4 3 Measurement (3) Starting the FDC NX600 Press the power button to start the FDC NX600. NOTE If the equipment is connected to DMS, switch on the equipment before starting DMS. Failure to observe this precaution could interfere with data transfer. IMPORTANT: 3 If the power has been shut off during measurement due to a power outage or other issue, replenish the tips (refer to Section 5.2) and replace the mixing cups (refer to Section 5.3). The analyzer may lose the information of the usage positions of the consumables and use wells (holes) of used mixing cups. Power button NOTE When the power switch is not in the ON position, push the switch to the " | " side. Power switch 0013 3-5 NOTE The listing to the left is displayed during start-up once a month. Perform the periodic maintenance listed in Chapter 6. 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement (4) Checking the date and time Check that the date and time displayed on the LCD are correct. If they are incorrect, set the date and time using [Date & Time settings] in Normal mode. è (Refer to Section 9.2.1) IMPORTANT: If the date and time are not adjusted correctly, the analyzer may fail to determine the expiration of the slides and calibration, and measurement results might be incorrect. (5) Loading consumables Load the consumables (tips, mixing cups, diluent, and reference fluid) in accordance with the type of measurement. è Refer to Chapter 5 (6) Reading a QC card QR code The FDC NX600 can compensate for the differences of the slide production lots by reading the QC card in the box of slides. When [Warming up] or [Ready to test] appears on the LCD, set the QR code on the left side, and then insert the QC card into the reader fully. è (Refer to Section 3.10) NOTE If the QC card is dirty or folded, the read error may occur. 897N203029A FDC NX600 Instruction Manual 2022.01 3-6 3 3 Measurement 3.3 Basic Measurements NOTE By default, this analyzer switches off the light source lamp to save lamp life when the analyzer has not been used for the time configured under [Lamp management] in Normal mode. When the lamp is off, refer to Section 3.3.1 before operating the equipment. WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. 3 (1) Checking the LCD When [Ready to test] appears on the LCD, make sure that the analyzer is not applying samples. NOTE You can stop sample application by pressing the [STOP] button. NOTE During sample application, a new sample for the next measurement cannot be loaded on the analyzer. (2) Setting tips RE Open the sampler cover, and then check the tip rack has been loaded with tips. When there are not enough tips, tap the tip display on the LCD to display the replacement screen, and then load the tips. For details of the loading method, refer to Section 5.2. IMPORTANT: If the power has been shut off during measurement due to a power outage or other issue, replenish the tips and replace the mixing cups (refer to Section 5.3). The analyzer may lose the information of the usage positions of the consumables and use wells (holes) of used mixing cups. Tips 3-7 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement (3) Loading samples and slides WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. 1. Open the packaging of the slide to be used. IMPORTANT: Take out the required number of slides from the refrigerator, and allow them to reach room temperature for 5 minutes or more with the condition of the individual package before unwrapping and using them. Otherwise, the measurement results and the analyzer operation may be affected. IMPORTANT: After the package has been unwrapped, finish measurement within 30 minutes. 2. Place slides to be measured in the slide cartridge with the test name facing up. Be sure to put the slide weight on top of the stacked slides. Slide weight NOTE The item to be measured is determined from the slide information on the back of the slide. Slide NOTE ISE slides can also be loaded at the same time as the slide cartridge. Including ISE slides, a maximum of 20 slides can be loaded. Slide cartridge 3. Load the slide cartridge while aligning its orientation. RE 897N203029A FDC NX600 Instruction Manual 2022.01 3-8 3 3 Measurement 4. Load the dedicated sample rack for the blood collection tube or FUJI sample tube to be used. NOTE To avoid a suction error or malfunction, use a sample tube that matches the label on the sample rack. NOTE Refer to Section 3.9 for sample tubes that can be used. 3 5. Remove the cap of the blood collection tube or FUJI Sample Tube that contains sample, and then load it on the dedicated sample rack. NOTE Make sure there are no bubbles in the sample in the blood collection tube or FUJI Sample Tube. Use a centrifuge to remove bubbles. If bubbles are present, the sample may be detected incorrectly. Blood collection tube or FUJI Sample Tube 3-9 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement (4) Registering sample information 1 2 3 4 5 6 7 Enter or select the sample information (sample number, sample ID, reference standard, and sample type). 1 Enter the sample number using the software keyboard (Section 2.3). 2 Obtain order information from the host PC via a DMS connection. è Refer to Section 4.6 3 Clear the entered information. 4 Enter the sample ID using the software keyboard (Section 2.3). 5 Select a reference standard from the list. 6 Select a sample type from the list. 7 Select a dilution factor from the list. è Refer to Section 3.6 and Section 9.3.14 NOTE When the analyzer is connected to DMS, sample information can be acquired automatically from the work list. è Refer to Section 4.6 IMPORTANT: The sample type of the sample, slide, and the sample type setting of the FDC NX600 must be same. Otherwise, an incorrect measurement result will be obtained. IMPORTANT: Be sure to enter the correct sample number and sample ID. IMPORTANT: When measuring a low concentration protein fluid (such as urine), some bubbles may appear during measurement in the tip. If bubbles have formed during measurement, perform the measurement again from the beginning. 897N203029A FDC NX600 Instruction Manual 2022.01 3-10 3 3 Measurement (5) Starting measurement Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 3 Tap the [START] button to start measurement. IMPORTANT: Close the disposal box. If it is open, the measurement results will be affected. 3-11 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 3 When the spotting operation is completed, the screen on the left is displayed and the time until the measurement is completed is displayed. Time until the end of measurement Measurement start time 897N203029A FDC NX600 Instruction Manual 2022.01 3-12 3 Measurement (6) Checking the measurement results After measurement, tap the button to check the test results. The test results are printed out in the order that the slides were loaded in the cartridge. To retest (Section 4.4), tap the 3 (retest) button. NOTE For the details of the measurement results screen, refer to Section 3.4. (7) Ending measurement To complete the measurement, unlock the sampler cover and open it, and then remove the measured sample. NOTE When a measurement is not being performed, empty the disposal box as appropriate. The disposal box can hold 100 slides and 50 AUTO TIPS. Always empty the disposal box before these numbers are exceeded. 3-13 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 3.3.1 Lamp Off and Measurement-start Scheduling By default, when this analyzer has not been used for some time (a time from 20 through 240 minutes can be set) while [Ready to test] is displayed on the LCD, it switches off the light source lamp to save lamp life. After the lamp is switched on again, the start of measurement can be scheduled even when [Lamp warming up] is displayed on the LCD. NOTE Do not use measurement-start scheduling for measurements using PF. (Refer to Section 4.3) After inversion, whole blood samples start to get separated. PF filtering may not occur normally, and an error may occur. NOTE The start of measurement cannot be scheduled in manual sample application mode. The touch panel screen becomes darker when the lamp is off. If the screen is tapped, the lamp switches on, and [Lamp warming up] is displayed on the LCD. In the case of the Measurement-preparation screen, the start of measurement is scheduled if the [START (Reserve)] button is tapped. If measurement is possible, it starts automatically. NOTE To cancel a scheduled measurement-start, tap the [STOP] button. 897N203029A FDC NX600 Instruction Manual 2022.01 3-14 3 3 Measurement 3.4 Measurement Results Screen After measurement has ended, tap then check the measurement results. button, and NOTE Results of the last 270 samples are backed up in data. Result data display area 3 3 < Display of measurement results > 1 Green: Within reference interval Red: Outside reference interval Or, when the measurement result is not displayed due to an error Lower limit of reference interval Upper limit of reference interval 1 2 Green: All measurement results are within reference interval. Red: The results include values outside the reference interval. Or, when an error has occurred. 2 3 Measurement results that include values outside the reference interval are indicated by , and the corresponding test name and value become red. If tapped, the details screen appears. will be indicated when an error has ocAnd, curred. will be indi* Also, in case of the following, cated. • When a colorimetric/ISE test result is outside of the measurement range (determination range) or the qualitative range • When a result of calculated parameter is displayed by [****] NOTE If measurement items that display H, L, #, or other characters are tapped, the details screen appears. 3-15 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement < Buttons of the measurement results screen > 1 2 1 button Close the measurement results screen, and then return to the Measurement-preparation screen. 2 Change sample No. button Change the sample number for which measurement results are to be displayed. 3 4 5 6 3 Feed button Feed the recording paper with a fixed amount. 4 Print button Print all measurement results of the displayed sample. 5 Send button 7 Send the measurement results of the displayed sample to a PC (via DMS). 6 Retest button 8 9 a 0 9 NOTE When the past test results are displayed, the results are displayed just as they were obtained using the settings when they were measured, such as Unit conversion, Correlation coefficients, Reference interval range, Test item name. The test results sent by the [Send] button have also the same format. Perform the measurement again under the same conditions for sample number, sample ID, reference standard, and sample type. Only the dilution factor can be changed. è Refer to Section 4.4 7 Result data scroll button Change the display of the measurement result data of the slide. 8 Sample No. tab Tap to display the measurement results of the selected sample number. 9 Sample No. tab scroll button Change the display of the Sample No. tab. This button is available when five or more samples have been measured. 0 Date scroll button Change the measurement date. a Display calendar button Display the calendar. If a date within the calendar is tapped, the measurement results of that day are displayed. è Refer to the next page. 897N203029A FDC NX600 Instruction Manual 2022.01 3-16 3 3 Measurement < Calendar display > 1 1 Measurement month scroll buttons Change the display of the calendar month. 2 Date buttons 2 Display the measurement results screen of a date. The currently displayed date is displayed in blue. NOTE Dates with no measurements are displayed in gray and cannot be selected. 3 3-17 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 3.5 ISE Tests IMPORTANT: Analyzer cannot detect expiries of ISE slides if the installed software is version 1.0. Be sure manually that slides are still valid. Otherwise, with the software version 1.1 or later is installed, analyzer can automatically detect expiries of the ISE slides that will expire 2023 or later or whose lot numbers are 130000 to 139999. NOTE When urine (U) is selected as the sample type for measurement, the dilution setting is set to 2x automatically. Therefore, diluent is needed in addition to reference fluid. The data is calculated automatically, so the results can be reported without adjustment. NOTE Refer to the “Instructions for Use” of the slide for usable diluent for urine measurements. NOTE Two tips are necessary for each slide; one for sample, the other for reference fluid. 1. Load the reference fluid. For the reference fluid container (FUJI plain tube 0.5 mL or 1.5 mL), use the container configured for [Tube setting for diluent & reference fluid] in Mode function for administrator (Section 9.3.15). For details of the loading method, refer to Section 5.5. 2. Check quantity of tips. Reference fluid • Loading tips è Refer to Section 5.2 3. Load diluent for urine sample measurement. Also, check the quantity of mixing cups. If the number is insufficient, replace with new ones. • Loading mixing cups è Refer to Section 5.3 • Loading diluent è Refer to Section 5.4 NOTE When performing an ISE measurement that requires dilution, use at least 3 tips and 1 well. 897N203029A FDC NX600 Instruction Manual 2022.01 3-18 3 3 Measurement 4. Check the sample type setting. Register sample information as necessary. è Refer to Section 3.3 (4) Sample type setting Remaining number of mixing cups 3 Arrows pointing right 矢印右向き 5. Load a sample and slides. Place the ISE slide in the slide cartridge with the spotting hole above and the arrows pointing right. For details of the loading method, see Section 3.3 (3). NOTE The direction of the arrow on the slide is opposite to the slide transport direction. NOTE When you measure an ISE slide and CM slide at the same time, we recommend loading the ISE slide last in the cartridge. è Refer to Section 4.5 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 3-19 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 7. Tap the [START] button to start measurement. 3 897N203029A FDC NX600 Instruction Manual 2022.01 3-20 3 Measurement 3.6 Dilution Tests There are 2 types of dilution measurement. • Dilute and measure automatically by using [Dilution setting] in Mode function for administrator to preset dilution factors for each test and sample type. è Refer to Section 9.3.14 • Dilute and measure all items to be measured by setting a dilution factor for each sample at the sample information registration screen. The second method at the sample information registration screen is explained in this section. IMPORTANT: 3 Bubbles may appear in the tip when diluting a plasma/serum sample more than 20 times or diluting a urine sample. If bubbles have formed during dilution, perform the measurement again from the beginning. NOTE The results of measurements that included dilution are calculated automatically, so the results can be reported without adjustment. NOTE The maximum number of dilution measurements from one mixing cup well (hole) is three. When the diluted sample in the well is insufficient, the analyzer automatically dilutes using the next well. NOTE Do not dilute whole blood samples. NOTE The default dilution factors designated by the manufacturer are as follows: ISE urine tests: 2x; CRP tests: 21x. Excluding ISE tests, the following section explains the relationships among the dilution factor configured for each sample at the sample information registration screen, the dilution factor for each test configured by [Dilution setting] in Mode function for administrator, and the dilution factor designated by the manufacturer. For ISE tests, the analyzer ignores the dilution factors configured for each sample or each test and using default dilution factor of the manufacturer. • Dilution factor when configured for each sample at the sample information registration screen Dilution factor configured for each sample × Default dilution factor • Dilution factor when not configured for each sample at the sample information registration screen (Configured by [Dilution setting] in Mode function for administrator) Dilution factor for each test × Default dilution factor (Example) With a CRP measurement with the dilution factor for each sample or test configured to 3x, the analyzer will automatically dilute 63x. NOTE Only one type of diluent can be used to measure a sample. 3-21 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement (Example) When diluting Sample 2x Diluent 1. Check the quantity of the consumables (tips and mixing cups). If consumables are insufficient, supplement or replace them. • Loading tips è Refer to Section 5.2 • Loading mixing cups è Refer to Section 5.3 2. Load the diluent. • Loading diluent è Refer to Section 5.4 RE 3 3. Load a sample and slides. For details of the sample and slide loading method, refer to Section 3.3 (3). 4. Tap the [Dilution] button. 897N203029A FDC NX600 Instruction Manual 2022.01 3-22 3 Measurement 5. In the list, tap [x2]. - : Dilution is performed according to [Dilution setting] in Mode function for administrator. None: No dilution NOTE If necessary, enter or choose other sample information. 3 3-23 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 3 7. Tap the [START] button to start measurement. 8. The dilution factor is added on the test result. 897N203029A FDC NX600 Instruction Manual 2022.01 3-24 3 Measurement 3.7 CRP Tests NOTE Each time a new box of CRP slides is used, the new QC card must be read. è Refer to Section 3.10 NOTE Calibration is necessary for CRP measurements under the following conditions. • When the slide lot has changed • When the validity term of calibration has expired (For the validity term, refer to the “Instructions for Use” of the slides) 3 NOTE If the validity term has expired, a “#” warning mark is printed along with the measurement results. NOTE Specified diluent and mixing cups are necessary for CRP measurements. NOTE When the results of a measurement will exceed the measurement range, dilution can be performed automatically at 21x or more by configuring the dilution factor at the sample information registration screen. (When 2x has been set at the sample information registration screen, the dilution factor is 42x.) Diluent for CRP 1. Check the quantity of the consumables (tips and mixing cups). If consumables are insufficient, supplement or replace them. • Loading tips è Refer to Section 5.2 • Loading mixing cups è Refer to Section 5.3 NOTE When performing measurements that require dilution, at least 2 tips are used. NOTE When performing CRP tests that require dilution at 2x or more, 2 wells (holes) of mixing cups are used. 2. Load the diluent for CRP. • Loading diluent è Refer to Section 5.4 3. Load a sample and slides. CRP CM slide 一般比色 図-4.9 3-25 For details of the sample and slide loading method, refer to Section 3.3 (3). NOTE When measuring a CRP slide and CM slide at the same time, stack the CRP slide last or first in the cartridge. If a CM slide is measured before or after a CRP slide, extra tips are used. 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 4. Set the sample type to [Plasma/Serum]. NOTE To further set the dilution factor to 3x for CRP measurements, tap the [Dilution] button, and then specify 3x. è Refer to Section 3.6 NOTE If necessary, enter or choose other sample information. 3 5. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 6. Tap the [START] button to start measurement. 897N203029A FDC NX600 Instruction Manual 2022.01 3-26 3 Measurement 3.8 Shutting-down the FDC NX600 Sample 1. Remove the samples. 2. Remove the tubes of diluent and reference fluid from the analyzer. Diluent Reference fluid 3 RE Sampler cover 3. Close the sampler cover. 4. Make sure that measurement and mode operation are not in progress. CAUTION: Do not switch off the power during measurement or mode operation. Otherwise, the analyzer may malfunction. If the power was shut off during measurement, load tips and other consumables, and then run the measurement again from the start. Always replace mixing cups for dilution measurements. 3-27 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 5. Press the power button. 3 Power button 6. Tap [OK]. CAUTION: Shutdown will be automatically performed after the process completed. Make sure NOT to turn the power switch off. CAUTION: Were you not able to operate on the screen, press the power button at front to turn off the power. NOTE When the analyzer will not be used for a long time, set the power switch to the [○] side. 897N203029A FDC NX600 Instruction Manual 2022.01 3-28 3 Measurement 7. Empty the disposal box. WARNING: Do not touch used slides and tips with your bare hands, as this may cause infection. If you accidentally touch an item, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: 3 Used slides and tips are infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. NOTE When disinfecting the disposal box, use ethyl alcohol for disinfection or a 0.5% sodium hypochlorite solution. When a sodium hypochlorite solution was used, wash the disposal box well and dry it before use. 3-29 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 3.9 Sample Tubes for the FDC NX600 The following sample tubes can be used on the FDC NX600. • • • • • • Ø16×100 mm blood collection tube (when using PF) • Ø13×100 mm blood collection tube (when using PF) • Ø13×75 mm blood collection tube (when using PF) Ø16×100 mm blood collection tube Ø13×100 mm blood collection tube Ø13×75 mm blood collection tube FUJI Sample Tube 1.5 mL FUJI Sample Tube 0.5 mL 3.9.1 Blood Collection Tubes (1) Size and effective liquid surface range of blood collection tubes that can be used Outer diameter of Ø16 indicates Ø15.0 to 16.6 mm and Ø13 indicates Ø12.0 to 13.3 mm. The following illustrations show the design suction ranges. If the sample volume is low, use a FUJI Sample Tube (1.5 mL or 0.5 mL). Ø16 or Ø13×100 mm blood collection tube 〈φ16またはφ13 x100mm採血管〉 10 mm 10mm Not measurable 測定不能範囲 Ø13×75 mm blood collection tube 〈φ13 x75mm採血管〉 10 mm 10mm Not measurable 測定不能範囲 Measurable 測定可能範囲 Measurable 測定可能範囲 75 mm 75mm 100 mm 100mm 30 mm 30mm Not測定不能範囲 measurable Not measurable 測定不能範囲 52 mm 52mm Sample volume confirmation holes (left and right) Load the blood collection tube with filled sample in the rack, so that you can see the actual sample through the holes. 897N203029A FDC NX600 Instruction Manual 2022.01 3-30 3 3 Measurement (2) Tube contents and inspection items that can be measured IMPORTANT: Refer to the “Instructions for Use” of the slides for usable blood collection tubes and observe the listed items. IMPORTANT: For contents, inspection items, cap colors, and other details, contact the manufacturer of the blood collection tube to be used. 3 The tube contents of a blood collection tube determine which inspection items can be measured. The following section provides information about general blood collection tube contents and the inspection items that can be measured. Heparin blood collection tubes • Inspection items: Applicable to all items. For GLU and NH3, measure the samples as soon as possible after blood collection. • General color of cap: Green (in Japanese market) Plain (coagulant included) blood collection tubes • Inspection items: All other than GLU and NH3 • General color of cap: Red (in Japanese market) NaF blood collection tubes • Inspection item: GLU (blood sugar) only • General color of cap: Gray (in Japanese market) EDTA blood collection tubes • Inspection items: NH3 only. After blood collection, keep samples in ice and measure the samples as soon as possible. • General color of cap: Pale purple (in Japanese market) 3-31 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement 3.9.2 FUJI Sample Tubes Use these tubes in the place of blood collection tubes when the sample volume is low. WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. 1.5 1.0 0.5 Upper 検体上限 limit of sample Lower limit 分離用 of 検体下限 sample for separation Make sure that the sample volume is sufficient by referring to the illustration to the left. The top surface of the sample fluid must be within the shaded section of the tubes in the illustration. 0.5 NOTE If the sample in the tube is below the suction limit mark, the analyzer will indicate an insufficient sample volume error. 0.3 0.1 Suction limit 吸引限界 0025 NOTE To aspire only the supernatant fluid (plasma) of the centrifuged sample (and avoid aspiration of red blood cells), at least 1.0 mL (for a 1.5 mL tube) or 0.3 mL (for a 0.5 mL tube) of a whole blood sample for separation is necessary. NOTE Only 5 tests or fewer can be performed using a 0.3 mL whole blood sample with a FUJI Sample Tube (0.5 mL). Effective liquid surface range of a FUJI Sample Tube Upper 検体上限 limit of sample 1.5 検体上限 0.5 1.0 0.3 0.1 0.5 吸引限界 注:検体は、50μl 以上必要です。 Suction limit 吸引限界 富士チューブ NOTE: At least 50 µl of sample is needed. サンプルカップ2ml 897N203029A FDC NX600 Instruction Manual 2022.01 3-32 3 3 Measurement 3.9.3 Assembling the Sample Rack Assemble the sample rack according to the type of blood collection tube and tube to be used. IMPORTANT: Incorrect assembly may cause bad suction, which may adversely affect the test results or cause a failure. <Assembly preparation> The following parts are included in the bundled items. 3 NOTE The number in parentheses is the number of parts included in the bundled items. NOTE The spacer (9) is are not used for the FDC NX600i. 2Tube holder (1) for φ 16mm tube 1Sample rack body (4) 4Tube holder (1) for FUJI tube 1.5mL 3Tube holder (1) for φ 13mm tube 5Tube holder (1) for FUJI tube 0.5mL 6Calibrator CP holder (1) 0Pin (9) H H aLabel (1) 7Spacer (1) for 75mm tube H = 54.5mm (long) 8Spacer (1) for 100mm tube H = 32mm (medium) H 9Spacer (1) for PF 75mm tube H = 25mm (short) Note: There are upper- and lower-sides of spacers. Do not turn them upside down. 3-33 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement <Sample rack variations> (a) Sample rack for φ16mm × 100mm Affix the label After setting the 8 spacer, insert the 2 tube holder Set the 8 spacer (H=32mm) (hole is down) Insert the 0 pin as shown 3 (b) Sample rack for φ13mm × 100mm Affix the label After setting the 8 spacer, insert the 3 tube holder Set the 8 spacer (H=32mm) (hole is down) 0 Pin (c) Sample rack for φ13mm × 75mm Affix the label After setting the 7 spacer, insert the 3 tube holder Set the 7 spacer (H=54.5mm) (hole is down) 0 Pin (d) Sample rack for FUJI sample tube 1.5mL Affix the label Insert the 4 tube holder 0 Pin 0 Pin 897N203029A FDC NX600 Instruction Manual 2022.01 3-34 3 Measurement (e) Sample rack for FUJI sample tube 0.5mL Affix the label Insert the 5 tube holder 0 Pin 3 (f) Sample rack for φ16mm × 100mm when using plasma filter PF Affix the label 0 Pin 0 Pin (g) Sample rack for φ13mm × 100mm when using plasma filter PF Affix the label 0 Pin 0 Pin (h) Sample rack for φ13mm × 75mm when using plasma filter PF Affix the label Set the 9 spacer (H=25mm) (hole is down) 0 Pin 0 Pin 3-35 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement (i) Sample rack for calibrator CP Affix the label Insert the 6 calibrator CP holder 0 Pin 0 Pin 3 897N203029A FDC NX600 Instruction Manual 2022.01 3-36 3 Measurement 3.10 QC Card System (1) Overview The QC card system compensates for variations between lots of FUJI DRI-CHEM slides and ensures uniform performance. Y Y 標Reference 準液基準値 liquid standard value Y 3 Y Calculation method A variety of the reference fluids developed for FUJI DRI-CHEM slides are measured on each lot of slides. The relationship between the actual measurement values X (x1, x2, x3) and the standard values Y (y1, y2, y3) is expressed by the formula Y=c+dX+eX2. The lot compensation coefficients (c, d, and e) derived from the formula are used to compensate for variations in X of each lot. yy33 yy22 yy11 0030 x11 x22 x33 XX スライド (補正前) の実測値 Actual measurement values of slideX(before compensation) X (2) Inputting lot compensation coefficients using a QC card The lot compensation coefficients are recorded on QC cards, and one QC card is packed with each box of slides. When data is input from a new card, the previously stored lot compensation coefficients are updated. Always read the new QC card when you switch to a new box of slides, because each QC card also contains slide data essential for measurement, in addition to the lot compensation coefficients. NOTE The “e” values recorded on the QC cards are expressed as e×103. NOTE The analyzer records the QC card information of up to 5 lots for each test. Reading QC cards of more than 5 lots results in deletion of QC information starting from the oldest. (3) Using QC cards A QC card must be read when a new test slide will be used or when changing to a new lot of slides. Read the QC card while [Ready to test] or [Warming up] is displayed on the LCD. 3-37 897N203029A FDC NX600 Instruction Manual 2022.01 3 Measurement Set the QR code on the left side, and then insert the QC card into the card reader fully. NOTE If the QC card is dirty or folded, the read error may occur. QR code 3 The slide test and production number (lot) are displayed on the LCD and output to the printer as well. NOTE Do not insert another QC card while the slide test and lot number are displayed. NOTE When the analyzer beeps and the test is not displayed, insert the QC card into the card reader again. (4) Using DI cards When you receive DI cards from FUJIFILM due to a change of the slide’s manufacturing process, read the DI cards according to the instructions. The reading method is the same as for QC cards. 897N203029A FDC NX600 Instruction Manual 2022.01 3-38 4 4.1 Applicable Usage Methods CRP Calibration 4.1.1 Calibration Measurement Calibrate when you start a new production number (lot) of CRP slides or when calibration has expired. WARNING: As calibrator CP is derived from blood, it should be handled in the same way as a patient sample. When handling calibrator CP, always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: 4 Perform the calibration measurement using Unit (A) (mg/dL). IMPORTANT: When the calibrator cap is opened, there may be a liquid film on the bottle opening. In this case, always remove the film before the calibration process. IMPORTANT: The cap of the calibrator should not be open more than 5 minutes during a single calibration. IMPORTANT: Use a calibrator 10 times or less, even during its validity term. IMPORTANT: If the date and time are not adjusted correctly, the analyzer may fail to determine the expiration of the slides and calibration, and measurement results might be incorrect. NOTE When performing calibration measurements, the analyzer automatically dilutes the calibrator. NOTE Let CRP diluent and calibrator stand until they reach room temperature before use. 897N203029A FDC NX600 Instruction Manual 2022.01 4-1 4 Applicable Usage Methods 1. Read the QC card via the analyzer. Read the QC card of the slide to be calibrated. è Refer to Section 3.10 2. Tap the button. 3. Tap the [Calibration] button. 4 4. Check the displayed calibrator concentration settings (CP1, CP2, and CP3). IMPORTANT: When the calibrator concentration values are incorrect, calibration is not performed normally. NOTE Refer to the “Instructions for Use” of the calibrator. NOTE You can change the calibrator concentration settings by tapping the [Settings] button. In such a case, the administrator password must be entered at the Administrator password entry screen. For the calibrator concentration settings, refer to Section 4.1.2. 4-2 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 5. Tap the [START Calibration] button. The guide for setting consumables such as calibrators is displayed. 4 6. Check the quantity of the consumables (tips and mixing cups). If consumables are insufficient, tap the [Set consumables] button to replenish or replace them. • Setting of tips è Refer to Section 5.2 • Setting of mixing cups è Refer to Section 5.3 7. Load the diluent for CRP. • Loading diluent è See Section 5.4 IMPORTANT: Always use diluent for CRP. 897N203029A FDC NX600 Instruction Manual 2022.01 4-3 4 Applicable Usage Methods 8. Prepare for measurement of calibrator CP1. NOTE A set of calibration requires 6 CRP slides in total. (a) Set the sample rack for the calibrator CP. NOTE Set the sample rack securely so that it does not rise up. (b) Insert 2 CRP slides into the slide cartridge and set them on the analyzer. CP sample rack 4 CP NOTE Set the slide cartridge securely so that it does not rise up. Otherwise, the slide reading error may occur. (c) Open the lid of the CP and place it in the sample rack. IMPORTANT: The CP should be open no more than 5 minutes during a single calibration. (d) Close the sampler cover, and then lock it with the sampler cover lock. Diluent for CRP IMPORTANT: Always lock the sampler cover with a key during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 4-4 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 9. Tap the [START] button. The CP1 calibration spotting screen is displayed. 4 10. After CP1 sampling, open the sampler cover, take CP1 out from the sample rack, and then firmly close the lid. The CP2 calibration measurement start screen is displayed. 897N203029A FDC NX600 Instruction Manual 2022.01 4-5 4 Applicable Usage Methods 11. Perform measurement of calibrator CP2. Perform (b) to (d) of Step 8, and then tap the [START] button. The CP2 calibration spotting screen is displayed. NOTE If a slide with a different lot number has been loaded, an error screen is displayed. 4 If the [OK] button is tapped, the original screen returns. Check the slide and continue the measurement. 4-6 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 12. After CP2 sampling, open the sampler cover, take CP2 out from the sample rack, and then firmly close the lid. The CP3 calibration start measurement screen is displayed. 13. Perform measurement of calibrator CP3. Perform (b) to (d) of Step 8, and then tap the [START] button. The CP3 calibration spotting screen is displayed. 4 897N203029A FDC NX600 Instruction Manual 2022.01 4-7 4 Applicable Usage Methods 14. After CP3 sampling, open the sampler cover, take CP3 out from the sample rack, and then firmly close the lid. The calibration measuring screen is displayed. 4 15. When measurement ends, the Calibration measurement result screen is displayed. NOTE Print the calibration measurement results and store them for the calibration validity term. To print, tap the button. NOTE The calibration parameters are saved to the analyzer automatically. 16. Tap the [OK] button to end calibration. 4-8 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods (Supplement) List display area The lot numbers (6 digits) and validity terms of read QC cards (CRP) are displayed in a list. Items are displayed in the order the QC card was read. Items within their validity term have their dates displayed in black. The following items are displayed in red: the date when the calibration validity term has expired, [Slide expired] when the validity term of the slide has expired, and [Unexecuted] when calibration is not been performed. 4 897N203029A FDC NX600 Instruction Manual 2022.01 4-9 4 Applicable Usage Methods 4.1.2 Configuring Calibrator Concentration Correct the calibrator concentration when the concentration included with the calibrator CP to be used and the concentration in the calibrator concentration display area at the Calibration Lot No. screen are different. 1. Tap the [Settings] button. The administrator password entry screen appears. 4 2. Enter the administrator password, and then tap the [OK] button. The Calibrator concentration configuration screen appears. NOTE The administrator password is necessary when moving from the Function menu screen to the Administrator function screen. 4-10 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 3. Tap the [Input] field of each CP. The software keyboard screen appears. 4 4. Enter a value, and then tap the [OK] button. For the details of the software keyboard screen, refer to Section 2.3. IMPORTANT: Enter the value using Unit (A) (mg/dL). 897N203029A FDC NX600 Instruction Manual 2022.01 4-11 4 Applicable Usage Methods 5. If the button is tapped, the entered values are registered, and the analyzer returns to the Calibration Lot No. screen. NOTE If the [Reset] button is tapped, the enbutton tered values are reset. If the is tapped, the entered values are registered and printed. 4 4-12 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 4.2 Manual Spotting (Colorimetric Slide) 4.2.1 Measuring by Manual Spotting (Colorimetric Slide) WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: In the case of a CM slide, use a dedicated pipette that matches the sample application volume specifications in the "Instructions for Use" of the slide. Carefully read the operation method of the pipette in Section 4.2.2. (Be aware that the sample application volume of GLU-W is 6 µL, which is different from other slide items.) NOTE Manual sample application is not possible with ISE slides. NOTE When the [Dilution] button was used to configure the dilution factor in manual sample application mode, the measurement value is calculated by a dilution factor. 1. Prepare the pipette, slides, and a sample. 2. Tap the 897N203029A FDC NX600 Instruction Manual 2022.01 button. 4-13 4 4 Applicable Usage Methods 3. Tap the [START Manual pipetting] button. 4 4. Register the measurement information. NOTE You can also display the measurement results screen from the screen on the left and request a retest. è Refer to Section 4.4 5. Load slides. For details of the loading method, see Section 3.3 (3). Manual pipette cover 6. Place the Manual pipette cover on the slide cartridge. NOTE If the cover is not set, the analyzer may fail to read the slide information and cannot start the measurements. RE 4-14 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 7. Aspirate the sample by using the pipette. NOTE Refer to the “Instructions for Use” of the slide for the amount of sample to be aspirated (applied). 8. Prepare sample pipetting. Hold the pipette and touch the tip mounted on the pipette to the guide (your side corner) of the Manual pipette cover. (This helps position the tip in the center of the slide.) 4 Guide the tip using your side corner. 9. Tap the [START] button. A slide is transferred to the sample application position. 10. Prepare so that sample application with the pipette is always possible, and then tap the [START] button. 897N203029A FDC NX600 Instruction Manual 2022.01 4-15 4 Applicable Usage Methods 11. After the progress bar has finished (all □ change to ■), apply the sample to the slide while [Spot now] is displayed. The spotted slide is automatically transferred to the measurement section after the display of [Spot now] disappears. IMPORTANT: Apply the sample while the buzzer is sounding (while [Spot now] is displayed). Otherwise, the measurement results will be inaccurate. IMPORTANT: 4 When applying sample to a CM slide, apply the sample within 1 mm of the center of the slide. Otherwise, the measurement results will be inaccurate. Run the measurement again. 参照液 搬送方向 (インキュベータ方向) Apply sample to the center スライドの中央に点着する of the slide (CM (比色スライ ド)slide). NOTE The LCD may not display [Wait] depending on the operation conditions. 12. To apply sample to the next slide, repeat the procedure from Step 5. 13. Press the [HOME] button to end manual sample application. NOTE The sampler unit moves. Apply sample to the center of the slide (CM slide) 4-16 897N203029A FDC NX600 Instruction Manual 2022.01 検体 4 Applicable Usage Methods 4.2.2 Using a Pipette WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: Mount the tip securely into the nozzle of the pipette. If mounting is poor, the tip may be dislodged, and the sample may scatter. 4 Applying samples to slides 1. Move the tip of the pipette near the slide. 2. Slowly push the push button and make a drop. 液滴を作ってスライドに近づけます 液滴を作ってスライドに近づけます Make 液滴を作ってスライ a drop and then move it near the slide ドに近づけます 3. Place the drop on the center of the slide. IMPORTANT: 液滴をスライドの中央に点着します 液滴をスライ します Place the drop ドの中央に点着 on the center of the slide 液滴をスライドの中央に点着します Always make a drop at the top of the tip, and then apply the sample to the slide. If the sample is discharged to the slide surface without making a drop, the measurement results may be affected. While the slide is absorbing the ながら スライドが検体を吸い取り sample, it diffuses 内部に拡散 していきinside ます 0057 0057 0057 スライドが検体を吸い取りながら 内部に拡散 していきます ながら スライ ドが検体を吸い取り 内部に拡散していきます 897N203029A FDC NX600 Instruction Manual 2022.01 4-17 4 Applicable Usage Methods 4.3 Tests Using a Plasma Filter (PF) By using a PF, you can filter a plasma sample from a whole blood sample without using a centrifuge. NOTE The FDC NX600i cannot measure using a PF. IMPORTANT: Read the “Instructions for Use” of the PF before measurement. IMPORTANT: Be sure to read the PF card when changing to a new production lot of PFs. CAUTION: Do not tap the [START] button when the sample has already been filtered by the PF. If the [START] button is tapped, the analyzer may malfunction. 4 Precautions for samples NOTE Use whole blood samples that have a hematocrit value (Hct) of 20 to 55%. Up to about 350 μL of plasma (including 85 μL of dead volume) can be filtered. The amount of filtered plasma decreases as the hematocrit value increases. In the case of a hematocrit value of 55%, about 185 μL of plasma (including 85 μL of dead volume) can be filtered. The amount of filtered plasma depends on the hematocrit value and viscosity of the blood. NOTE Hemolyzed samples provide inaccurate results. NOTE If blood cells in a sample have precipitated, mix the sample, place it in a sample rack, and then perform measurement immediately. Otherwise, the amount of filtered plasma may be insufficient. Precaution for blood collection tubes IMPORTANT: Use the blood collection tubes listed in Section 4.3.2. 4-18 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods PF cards (1) Overview When the measurement values for filtered plasma are compared to the values for plasma separated on a centrifuge, systematic differences may be found. Always read the PF card before using a PF. The PF card contains information to compensate for these differences. (2) Using PF cards One PF card is included in one PF box (with 50 filters). Read a PF card in the same way a QC card is read. Be sure to read the PF card when changing to a new box of PF. The imported production number (lot) is displayed on the Measurement-preparation screen. Make sure that this number matches the production number (lot) of the PF in use. (è Refer to Section 2.2.1) NOTE A PF card cannot be read during measurement. 4 897N203029A FDC NX600 Instruction Manual 2022.01 4-19 4 Applicable Usage Methods 4.3.1 Measurement 1. Prepare the slides. 2. Prepare the consumables. Prepare the tips, mixing cups, diluent, and reference fluid as required. • Loading tips è Refer to Section 5.2 • Loading mixing cups è Refer to Section 5.3 • Loading diluent è Refer to Section 5.4 • Loading reference fluid è Refer to Section 5.5 PF sample rack 4 4-20 3. Load the sample rack for PF and the slide to be measured. 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 4. Load the sample. (1) Place the specified amount of whole blood sample in a blood collection tube. NOTE The PF has about 3 mL of dead volume (that will be absorbed into the filter). If the sample volume is low, the filtered sample volume will be too low for measurement. Furthermore, sample that was aspirated into the PF filter cannot be reused. NOTE If the sample volume is too high, the sample may overflow from the blood collection tube when the PF is loaded. There must be empty space in the tube. (Refer to Section 4.3.2.) (2) Mix the whole blood sample in the blood collection tube by turning it over five or six times. NOTE Make sure that bubbles do not form. (3) Remove the cap from the blood collection tube, and then put it on the PF sample rack. (4) Insert a PF into the blood collection tube. Make sure the PF does not rise from the sample rack. 5. Register the measurement information. IMPORTANT: Tests using a PF measure filtered plasma, so always configure the sample type to [Plasma/Serum]. 897N203029A FDC NX600 Instruction Manual 2022.01 4-21 4 4 Applicable Usage Methods 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 4 7. Tap the [START] button to start measurement. 8. After measurement is ended, discard the used PF, and then remove the sample container. WARNING: When discarding used PF, always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If any part of the body comes in contact with a sample, immediately rinse the contaminated body part thoroughly under running water and then use ethyl alcohol as a disinfectant. Seek medical assistance if necessary. WARNING: Used PF is infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. 4-22 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 4.3.2 Blood Collection Tubes for Tests Using a PF NOTE The FDC NX600i cannot measure using a PF. The following blood collection tubes can be used with PF measurements: φ16×100 mm, φ13×100 mm, and φ13×75 mm. (Outer diameter of Ø16 indicates Ø15.0 to 16.6 mm and Ø13 indicates Ø12.0 to 13.3 mm.) IMPORTANT: Use blood collection tubes containing Heparin. IMPORTANT: Refer to the “Instructions for Use” of the plasma filter (PF) and always follow the instructions. NOTE Do not use blood collection tubes that may cause the tip of the sample suction nozzle of the PF to be physically blocked by additives or other materials. NOTE As shown in the illustration below, the required sample volume is 6.5 mL or more for Ø16×100 mm tubes, 6 mL or more for Ø13×100 mm tubes, and 3 mL or more for Ø13×75 mm tubes. PF PF 10 mm of empty 10mmの空き space is necessary スペースが必要です 1010mmの空き mm of empty space is necessary スペースが必要です Measurable 測定可能範囲 Measurable 測定可能範囲 75 mm 75mm 100 mm 100mm 検体は The required sample volume φ16x100mm管で6.5ml以上 is 6.5 mL or more for Ø16×100 φ13x100mm管で6ml以上 mm tubes, and 6 mL or more for必要です Ø13×100 mm tubes. 検体は The required sample φ13x75mm管で3ml以上 volume is 3 mL or more for 必要です Ø13×75 mm tubes. 0023 NOTE Use PF sample racks with the following labels for tests using a PF. 897N203029A FDC NX600 Instruction Manual 2022.01 4-23 4 4 Applicable Usage Methods 4.4 Retesting You can test again using the previous sample number, sample ID, sample type setting, and reference interval setting by tapping the (retest) button at the measurement results screen. Furthermore, when a measurement cannot be performed due to an error, you can skip the re-entry and reconfiguration of the setting information by tapping the (retest) button. When a PF measurement is run again, the analyzer does not filter using the PF, but instead uses plasma that was previously filtered by the PF. IMPORTANT: The (retest) button is designed only to automatically enter the sample number and sample ID of the previously measured sample. Do not use this button to retest samples from before the previous sample. CAUTION: 4 Do not tap the [START] button when the sample has already been filtered by the PF. If the [START] button is tapped, the analyzer may malfunction. Retesting after filtering by the PF • When you will rerun a measurement using the PF, do not use the [START] button to restart the measurement. Always use the (retest) button. • Make sure that the sample volume is sufficient. • Make sure that the liquid surface of the filtered sample is level. IMPORTANT: Do not retest GPT (ALT) of PF filtered plasma because its activity decreases over time. 4-24 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods < How to retest > 1. When [Ready to test] appears on the LCD, make sure that the analyzer is not applying samples, and then load the consumables and sample. è Refer to Section 3.3 NOTE If the amount of remaining sample is insufficient, run the test with manual pipetting. è Refer to Section 4.2 2. Tap the (retest) button on the measurement results screen screen. NOTE The sample number or sample ID that was measured just before is displayed. NOTE In case of a sample scheduled for retesting, is displayed. NOTE Tap the [Dilution] button to set the dilution factor. 897N203029A FDC NX600 Instruction Manual 2022.01 4-25 4 4 Applicable Usage Methods 3. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 4 4. Tap the [START] button to start retesting. NOTE When a test is rerun, PF filtering process is not performed. 4-26 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 4.5 Slide Loading Order When Measuring Different Items Simultaneously NOTE Any order of slide stacking can be measured. However, please follow the recommended slide stacking rules below. Otherwise, the analyzer may consume excessive amounts of disposal tips and cups. (1) Measuring CRP and CM (colorimetric) slides CRP Always put a CRP slide first or last. If a CRP slide is loaded between CM slides, an extra tip will be used. CM slide 一般比色 図-4.5.2-1 4 (2) Measuring CM and ISE slides ISE 電解質 CM slide 一般比色 We recommend loading an ISE slide last. If an ISE slide is measured first, it might take extra time for measurement. If a CM slide is loaded after an ISE slide, an extra tip will be used. 図-4.5.2-1 (3) Measuring CRP, CM, and ISE slides Stack CM slides before CRP and ISE slides. ISE 電解質 CRP CM slide 一般比色 図-4.5.2-4 (4) Measuring CM slides that require dilution 5x dilution 5倍希釈 2x dilution 2倍希釈 図-4.5.2-2 If slides are configured with different dilution rates, stack the slides so that slides with same dilution factor are measured successively. Otherwise, extra tips and mixing cups will be used. NOTE If all CM slides have the same dilution factor, there are no restrictions on the stacking order. <Not recommended slide loading order> CRP 電解質 CM (Dil) CRP CM slide 一般比色 図-4.5.2-4 Only one loading position is prepared for the diluent in this analyzer. In the example, diluent (purified water etc.) for the colorimetric slide that requires dilution [CM (Dil)] is loaded and the measurement will start. The next measurement requires the specified diluent for CRP, so that an error (E0160 Incorrect diluent type error) occurs before the measurement, and the measurement will be completed. Rerun the CRP measurement from the beginning. 897N203029A FDC NX600 Instruction Manual 2022.01 4-27 4 Applicable Usage Methods 4.6 Automatic Acquisition of Sample Information (Work Lists) If the analyzer is connected to DMS, you can receive patient IDs and their sample information by button at the Measurement-preparation screen before measurement. tapping the NOTE DMS is sold separately. 1. Tap the button. The analyzer receives a work list (a list of patients for the samples to be measured) from DMS. 4 The Work list index screen appears. NOTE If there are no measurement requests, the buzzer sounds, and the message [There is no request information] is displayed. NOTE The display of the list is configured by [Work list selection setting] in Mode function for administrator. è Refer to Section 9.3.8 NOTE On DMS, "Sample No." is "Sample ID" and "Sample ID" is "Patient ID”. NOTE When there are multiple pages in a list, you can scroll the pages using the / buttons. NOTE If the [Cancel Work list] button is tapped, the analyzer returns to the Measurement preparation screen. 2. Tap the sample information you want to register. 4-28 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods The work list measurement items of the registered sample information are displayed. NOTE If the button is tapped, registration is cleared. NOTE The work list measurement items can be printed by tapping the button. 4 3. Tap the [OK] button to finish registration. The Measurement preparation screen (top screen) returns. NOTE The work list button lights. 897N203029A FDC NX600 Instruction Manual 2022.01 4-29 4 Applicable Usage Methods 4.7 Editing Sample Numbers and Sample IDs You can edit the sample number and sample ID of measurement results. You can also send edited data to a host computer. 1. To edit the sample number, tap the [No.] button at the measurement results screen (Refer to Section 3.4). 4 2. Enter the administrator password, and then tap the [OK] button. The Sample number entry screen appears. 3. Edit the sample number, and then tap the [OK] button. The analyzer returns to the measurement results screen. NOTE You can also edit the sample ID using the same procedure. IMPORTANT: Be sure not to make any entry errors. NOTE The sample ID can also be edited by using the barcode reader. 4-30 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods 4.8 Using the Sample Barcode Reader (Sold Separately) You can easily enter the sample ID and operator ID by using the optional sample barcode reader. IMPORTANT: If the sample barcode reader malfunctions or the print quality of the barcode label is poor, incorrect barcode data may be read. Make sure that the barcode data (sample ID) printed in the measurement results are correct. NOTE Carefully read the instructions included with the sample barcode reader before using it. A maximum of 13 alphanumerical characters can be input for a sample ID. Always use the sample barcode reader under lighting at 3,000 lux or lower and avoid direct sunlight. NOTE Always connect the cable of the sample barcode reader before switching on the FDC NX600. <When inputting the sample ID> 1. Tap the [Sample ID] field. The software keyboard appears. NOTE In the following cases, this operation is not required. The barcode reading is possible on the top screen (left): • When the USM function is enabled. • When the setting "Use operator ID" is OFF (USM function is disabled). è Refer to Section 9.3.19. NOTE You can also enter the sample ID using the barcode reader on the measurement results screen. 897N203029A FDC NX600 Instruction Manual 2022.01 4-31 4 4 Applicable Usage Methods 2. Read the barcode of the sample using the barcode reader. After reading is complete, the analyzer makes a beep sound. The sample ID is displayed on the LCD and is registered. バーコードを読み取ります NOTE Position the sample barcode reader properly facing to the barcode label. Otherwise, an incorrect reading may occur. NOTE Make sure the displayed ID is correct. NOTE When reading a barcode again, tap the [Sample ID] field, and then re-read the barcode. 4 4-32 897N203029A FDC NX600 Instruction Manual 2022.01 4 Applicable Usage Methods <When inputting the operator ID> 1. Tap the operator ID button. The software keyboard appears. 4 2. Read the operator ID using the sample barcode reader. After reading is complete, the analyzer makes a beep sound. The operator ID is displayed on the LCD and is registered. NOTE Position the sample barcode reader properly facing to the barcode label. Otherwise, an incorrect reading may occur. NOTE Make sure the displayed ID is correct. NOTE To re-input a barcode, tap the Operator ID button, and then re-read the barcode. 897N203029A FDC NX600 Instruction Manual 2022.01 4-33 5 5.1 Loading Consumables Loading Consumables To load tips and mixing cups, tap the display of consumables to display the loading screen. When loading is complete, press the [Finish] button to exit. NOTE If there are not enough consumables (tips, mixing cups, diluent, reference fluid, etc.), sample application may be stopped. NOTE Load consumables while the analyzer is not applying samples. 5 897N203029A FDC NX600 Instruction Manual 2022.01 5-1 5 Loading Consumables 5.2 Setting Tips Tips are required for all measurements. Check the number of remaining tips on the LCD before performing measurements. If necessary, add new tips. WARNING: During sample application, tap the [STOP] button to stop application, and then load the tips. Loading is unsafe during sample application because the sampler and the consumable disk are moving. 1. With [Warming up], [Lamp warming up] or [Ready to test] displayed on the LCD, make sure that the analyzer is not applying samples. 2. Tap the consumables display on the LCD. 5 Consumable display 3. The Consumables loading screen is displayed. 5-2 897N203029A FDC NX600 Instruction Manual 2022.01 5 Loading Consumables 4. Remove the tip rack from the consumable disk. NOTE Always remove the tip rack from the consumable disk before loading tips. RE Tip rack 5. Load tips in all holes of the tip rack. NOTE Always set tips in all holes of the tip rack. 5 6. Load the tip rack on the consumable disk. RE Tip rack loading position 897N203029A FDC NX600 Instruction Manual 2022.01 5-3 5 Loading Consumables 7. Check that tips have been set in all holes of the tip rack. And then, tap the [Replenish] button of tips. 8. Tap the [Finish] button to exit. 5 5-4 897N203029A FDC NX600 Instruction Manual 2022.01 5 Loading Consumables 5.3 Loading Mixing Cups Mixing cups are necessary for measurements requiring dilution (refer to Section 3.6) and CRP measurements (refer to Section 3.7). Furthermore, mixing cups are also necessary when urine (U) is selected for the sample type during an ISE measurement because measurement is performed automatically at 2x dilution. (Refer to Section 3.5) Check the quantity of mixing cups on the LCD before measurement. If necessary, load new mixing cups. Each measurement uses one mixing cup. If no mixing cups remain, a measurement cannot be run. WARNING: During sample application, tap the [STOP] button to stop application, and then load the mixing cups. Loading is unsafe during sample application because the sampler and the consumable disk are moving. WARNING: Always wear gloves when handling used mixing cups to avoid biohazards. NOTE The quantity of mixing cups that can be used is displayed on the LCD. When the amount remaining is insufficient, a dilution measurement cannot be performed. 1. With [Warming up] or [Ready to test] displayed on the LCD, make sure that the analyzer is not applying samples. 2. Tap the consumables display on the LCD. Consumable display 3. The Consumables loading screen is displayed. 897N203029A FDC NX600 Instruction Manual 2022.01 5-5 5 5 Loading Consumables Mixing cups 4. Remove the mixing cups from the consumable disk, and then set new mixing cups. WARNING: RE Used mixing cups are infectious industrial waste. Process this waste in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization, or disinfection. 5. Tap the [Replace] button of mixing cups. 6. Tap the [Finish] button to exit. 5 5-6 897N203029A FDC NX600 Instruction Manual 2022.01 5 Loading Consumables 5.4 Loading Diluent Diluent is necessary for dilution measurements (refer to Section 3.6) and CRP measurements (refer to Section 3.7). Dilution and CRP measurements each require a specified diluent. Before measurement, refer to the technical documentation for the diluent to be used. Furthermore, mixing cups are also necessary when urine (U) is selected for the sample type during an ISE measurement because measurement is performed automatically at 2x dilution. (Refer to Section 3.5) WARNING: During sample application, tap the [STOP] button to stop application, and then load diluent. Loading is unsafe during sample application because the sampler and the consumable disk are moving. IMPORTANT: Use specified tube for the diluent. Otherwise, measurement errors or malfunctions may occur. The specified tubes for diluent are the FUJI PLAIN TUBE (1.5 ml and 0.5 ml). IMPORTANT: The diluent for CRP measurements evaporates progressively, and the concentration of the fluid may affect the measurement values. Replace diluent for each test. NOTE For each CRP measurement, 0.1 mL of diluent is necessary. NOTE Configure the type of diluent container to be used via [Tube setting for diluent & reference fluid] in Mode function for administrator (refer to Section 9.3.15). FUJI PLAIN TUBE (1.5 ml) FUJI PLAIN TUBE (0.5 ml) 1.5 0.5 Suction range 1.0 Pour diluent up to this range 0.3 0.1 1. Prepare the diluent. Suction range 0.5 897N203029A FDC NX600 Instruction Manual 2022.01 Pour the following amounts into each type of FUJI PLAIN TUBE. For 1.5 mL tube: 1.0 to 1.5 mL For 0.5 mL tube: 0.3 to 0.5 mL NOTE The suction range is as shown on the left. When 0.5 mL of diluent was poured into a FUJI PLAIN TUBE of 1.5 mL, one dilution is possible for a CRP measurement. 5-7 5 5 Loading Consumables 2. With [Warming up] or [Ready to test] displayed on the LCD, make sure that the analyzer is not applying samples. Diluent 3. Load a tube containing diluent. RE 5 NOTE Tap the consumables display on the top screen and select the [Diluent] tab to display the diluent setting guidance screen. This operation is not required because only the guidance screen is displayed. 5-8 897N203029A FDC NX600 Instruction Manual 2022.01 5 Loading Consumables 5.5 Loading Reference Fluid Reference fluid is necessary for ISE measurements. (Refer to Section 3.5). WARNING: When handling reference fluid, follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the medical handling rules of your facility. If you accidentally touch an item, immediately rinse the contaminated body part thoroughly under running water. WARNING: Discard reference fluid according to the medical handling rules of your facility and related laws. WARNING: During sample application, tap the [STOP] button to stop application, and then load the reference fluid. Loading is unsafe during sample application because the sampler and the consumable disk are moving. IMPORTANT: Use a FUJI PLAIN TUBE (0.5 or 1.5 mL) for reference fluid. Otherwise, measurement errors or malfunctions may occur. Configure the type of container to be used via [Tube setting for diluent & reference fluid] in Mode function for administrator (Refer to Section 9.3.15). IMPORTANT: Replace FUJI PLAIN TUBES (1.5 mL) within 24 hours. Replace FUJI PLAIN TUBES (0.5 mL) for each measurement. FUJI PLAIN TUBE (1.5 ml) 1.0 FUJI PLAIN TUBE (0.5 ml) Pour reference fluid within the range. 1. Prepare reference fluid. Pour the following amounts into each type of FUJI PLAIN TUBE. For 1.5 mL tube: 1.0-1.5 mL For 0.5 mL tube: 0.3-0.5 mL 0.5 0.3 NOTE Use reference fluid after it reaches room temperature. 0.1 0.5 2. With [Warming up] or [Ready to test] displayed on the LCD, make sure that the analyzer is not applying samples. 897N203029A FDC NX600 Instruction Manual 2022.01 5-9 5 5 Loading Consumables Reference fluid 3. Load the tube containing the reference fluid. Press the lever to open the lid and load the container. IMPORTANT: Clean the cap of the reference fluid every time you load the reference fluid. è Refer to Section 6.8 NOTE Make sure the reference fluid cap is closed properly. 5 RE Reference fluid NOTE Tap the consumables display on the top screen and select the [Reference fluid] tab to display the reference fluid setting guidance screen. This operation is not required because only the guidance screen is displayed. 5-10 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6 Maintenance To keep the performance of the FDC NX600 at its best, users must perform daily and periodic maintenance, and designated repair personnel must perform maintenance. When repairs are necessary, contact repair personnel. WARNING: When you clean the analyzer, always wear gloves, a lab coat, and safety goggles. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Always unplug the power cord before cleaning. (Cleaning of the transfer bar and cleaning around the PF unit are excluded.) WARNING: Process contaminated swabs or cloths used for cleaning the analyzer as infectious waste. CAUTION: Before cleaning and maintaining the analyzer, get training from our technical support staff. 6 CAUTION: Refer to each section for information about usable solvents for cleaning and decontamination. CAUTION: Do not use alcohol to clean the sampler cover (translucent). Otherwise, the surface will be damaged. IMPORTANT: After cleaning, always connect the power cord, switch on the power, and then check operation. IMPORTANT: If you do not perform the maintenance in this manual, the analyzer’s performance and specifications will not be maintained, which may negatively affect the measurement values. IMPORTANT: Be sure to reassemble the parts removed for maintenance and tighten thumbscrews securely. Otherwise, measurement values may be negatively affected. NOTE If the LCD (touch panel section) becomes soiled, use a soft cloth or a cloth soaked in a small amount of ethyl alcohol to lightly wipe it. NOTE When cleaning the outer covers of the equipment, wipe using a soft cloth moistened with water or ethyl alcohol for disinfection (excluding the sampler cover). 897N203029A FDC NX600 Instruction Manual 2022.01 6-1 6 Maintenance 6.1 Periodic Maintenance Items 6.1.1 Items for Daily and Periodic Inspection by the User Users must perform the daily and periodic maintenance and replacement of parts in the following table. Location Cleaning interval Replacement interval Procedure Air filter Once a month - - Refer to Section 6.2. Transfer bar Once every 3 months, or as appropriate, such as when inaccurate measurement values occur - - Refer to Section 6.3. Incubator Once every 3 months, or as appropriate, such as when inaccurate measurement values occur - - Refer to Section 6.3. - - Refer to Section 6.3. Once every 3 months, or as appropriate, such as when inaccurate measurement values occur - - Refer to Section 6.3. Recording paper - - Appearance of red lines on both sides of paper Refer to Section 6.4. Light source lamp - - When a lamp replacement error Refer to occurs, or when the cumulative Section 6.5. lighting time exceeds 1,000 hours As appropriate, such as when a slide transfer Sample application unit error or tip ejection error occurs 6 Inspection interval ISE unit Sampler O-ring Once a week Slide reading section When slide reading errors occur frequently - Reference fluid cap packing When replacing reference fluid - PF sensor As appropriate, such as when errors related to PF extraction occur Once a month PF reference plate PF Suction Pad PF Tube - Once a month - Refer to Section 6.6. Once a year Once every 3 months Once every 2 years Once every 2 years or when the power has been off for one month Refer to Section 6.7. Refer to Section 6.8. Refer to Section 6.9. Refer to Section 6.9.4. WARNING: After inspection, always connect the power cord, switch on the power, and then check operation. 6-2 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.1.2 Items for Maintenance and Inspection by Repair Personnel Designated repair personnel must perform the maintenance and replacement of parts in the following table. Parts Replacement interval Interference filter Once every 4 years Syringe Once every 3 years Incubator pressure plate Inspection once every 3 years WARNING: After inspection, always connect the power cord, switch on the power, and then check operation. NOTE The standard number of slides used per year for this analyzer is assumed to be 7,200 colorimetric slides and 240 ISE slides. (In the case of 240 days of use per year, 30 colorimetric slides/day and one ISE slide/day) 6 897N203029A FDC NX600 Instruction Manual 2022.01 6-3 6 Maintenance 6.2 Cleaning the Air Filters The air filters should be checked and cleaned once a month. IMPORTANT: If the air filters are not cleaned, an increase in the temperature inside the analyzer may negatively affect the measurement values. NOTE Clean all 4 filters. 1. Switch off the analyzer power. 2. Remove the four louvers and pull out the air filters located inside of the louvers. Lift the louver upward 1 and then pull forward 2 to remove. Louver Louver 6 2 1 6-4 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 3. Remove dust adhering to the filter with a vacuum cleaner or running water. Air filter NOTE If you wash the filter with running water, make sure it is dried well before setting in the analyzer. 4. Insert the filter into the louver and put the louver back in the analyzer. Push up the louver 1 and then push it toward 2. 1 IMPORTANT: 2 If the analyzer is used without setting the filters, measurement values may be negatively affected. 897N203029A FDC NX600 Instruction Manual 2022.01 6-5 6 6 Maintenance 6.3 Inspecting and Cleaning the Transfer Bar, Incubator, Sample Application Unit, and ISE Unit Measuring results are affected by stains (from blood samples, etc.) inside of the incubator and ISE unit. These parts should be checked and cleaned at least once every three months. When a slide transfer error or a tip ejecting error occurred, clean the sample application unit. WARNING: Always wear gloves, a lab coat, and safety glasses while working. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Dispose of contaminants used for cleaning as infectious waste. IMPORTANT: Staining of the incubator, spotting parts, and electrolyte parts with samples such as blood may affect the measured values. 6 6-6 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.3.1 Reference Plate Level Check Perform [Reference plate level check] before cleaning. 1. Tap the button. 2. Tap the [Normal mode] button. 6 897N203029A FDC NX600 Instruction Manual 2022.01 6-7 6 Maintenance 3. Tap the [Reference plate level check] button to run a check. If check ends, the Check results screen appears. 4. Check the results of [Reference plate level check]. When has appeared in the check results, cleaning is necessary. Tap the [OK] button. Using the procedure on the following pages, clean the transfer bar, incubator, sample application unit, and ISE unit. In the case of Level 1 soiling (0-49%) In the case of Level 2 soiling (50-79%) In the case of Level 3 soiling (80% or higher) 6 6-8 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.3.2 Cleaning the Transfer Bar, Incubator, Sample Application Unit, and ISE Unit (1) Cleaning the transfer bar 1. Tap the [Cleaning mode] button in Normal mode. 2. Remove the slide cartridge and the tip rack from the analyzer. 3. Tap the [Start Cleaning] button. NOTE Keep your fingers clear as the analyzer runs. 897N203029A FDC NX600 Instruction Manual 2022.01 6-9 6 6 Maintenance 4. The screen to the left appears. 5. Loosen the 2 thumbscrews and pull the sample application unit cover frontward to remove it. 6 Thumbscrew 6. Remove the two slide transfer weights. NOTE There are 2 types of slide transfer weight, one each for CM tests and ISE tests. The slide transfer weight for CM tests has hooks on the left, so lift up from the right before removing it. When installing it, insert the hooks before loading it. For CM Slide transfer weight For ISE 7. Use gauze or a cotton swab that has been soaked in lukewarm or cold water to wipe the staining of the removed slide transfer weight. 6-10 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance Transfer bar 8. Use gauze or a cotton swab that has been soaked in lukewarm or cold water to wipe the upper surface of the transfer bar visible in the sample application unit. 9. Next, clean the incubator, sample application unit, and ISE test unit. NOTE To quit the cleaning, attach the removed slide transfer weight and sample application unit cover at their original positions, and then tap the button to return to the previous screen. 6 CAUTION: Keep your fingers clear as the analyzer runs. 10. Tap the power. button to switch off the CAUTION: Keep your fingers clear as the analyzer runs. 897N203029A FDC NX600 Instruction Manual 2022.01 6-11 6 Maintenance WARNING: Always unplug the power cord before the following procedure. (2) Incubator 1. Open the sampler cover and panel unit. Remove the panel unit pedestal. Panel unit pedestal 6 Connector Incubator cover 6-12 2. Unplug the connector of the incubator cable (low end) by turning it to the left. 3. Remove the incubator cover. 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 2 5 4 4. Loosen the 5 screws on the upper surface of the incubator, and then remove the incubator. 3 1 Thumbscrew 5. Moisten the cloth with lukewarm water. IMPORTANT: Wring the cloth when it is soaked too much. If water gets inside the machine, it will cause rust. 6. Place the removed incubator on a flat surface with the pressure plates facing up. IMPORTANT: Put on gloves before beginning wiping. Incubator pressure plate 7. If dirt is noticeable or a transport error occurs within the cleaning cycle, shorten the cleaning cycle (to at least once a week). 897N203029A FDC NX600 Instruction Manual 2022.01 6-13 6 6 Maintenance Tapered part 8. Wipe the tapered part of each pressure plate (the metallic luster taper part around the black part). NOTE Wipe the entire circumference as blood stains may be difficult to see by eye. 9. Wipe the surface (black part) of each pressure plate. Surface NOTE Wipe gently to avoid scratching, as it will affect the measurement values. IMPORTANT: 6 When wiping, change the cloth surface for each cell. Otherwise, dirt may be transferred. IMPORTANT: Do not use a cotton swab. The components of the cotton swab may negatively affect the measurement values. IMPORTANT: After cleaning, do not touch the pressure plate surface with your fingers. This may affect the measurement values. 10. Wipe the lower part (measurement cells) in the same way. Measurement cell 6-14 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 11. Loosen the 2 screws fixing the reference plate and remove the plate from the analyzer. Thumbscrew Reference black plate Reference plate Reference white plate Photometer 12. Turn over the reference plate and clean the white plate and black plate with a dry cotton swab. 13. Clean the photometer head with a dry cotton swab. 14. Load the reference plate. Tighten the 2 thumbscrews evenly. IMPORTANT: Tighten the two thumbscrews securely. Otherwise, the measurement results may be negatively affected. Thumbscrew Reference plate 897N203029A FDC NX600 Instruction Manual 2022.01 6-15 6 6 Maintenance (3) Sample application unit 1. Make sure that the sample application unit cover and two slide transfer weights were removed for cleaning of the transfer bar. Sample application unit 6 2. Wipe the sample application unit with gauze or a cotton swab moistened with lukewarm or cold water. (4) ISE unit probe Make sure there is no dust on the six ISE probes. When you clean the ISE probes, wipe off the dust using a dry cotton swab. NOTE Do not use solvents, such as methyl alcohol or isopropyl alcohol. Probes CAUTION: ISE probes are needle-shaped parts. Do not touch them directly by hand. Be careful not to bend the probes. 6-16 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.3.3 Reassembling Parts and Ending Cleaning 1. Reassemble the incubator. 2 Align the arrow marks on the incubator and the reference plate, and then tighten the 5 thumbscrews evenly by tightening opposing thumbscrews in numerical order (1 to 5). 5 4 3 1 IMPORTANT: Thumbscrew 4 Incubator cable 3 Incubator cover 1 Slide transfer anel unit 5P weight pedestal 2 Sample application unit Incubator cable Tighten the five thumbscrews securely. Otherwise, the measurement results may be negatively affected. 2. Reassemble each part as shown in the illustration on the left. 1 Load the slide transfer weights. 2 Load the sample application unit cover and tighten the thumbscrews. 3 Load the incubator cover. 4 While aligned with the groove, plug the connector of the incubator cable into the incubator and then lock it by turning it to the right. 5 Load the panel unit pedestal. Incubator cover IMPORTANT: Tighten the thumbscrews securely. Otherwise, the measurement results may be negatively affected. IMPORTANT: Be sure to attach the incubator cover. Otherwise, the measurement results may be negatively affected. NOTE When you insert the connector, make sure not to twist or apply too much pressure to the cable. If the cable is twisted more than once, it may break. 3. Close the panel section to end cleaning. Run [Reference plate level check] again, and then check the results. è Refer to Section 6.3.1 NOTE Wait until the temperature of the incubator is steady. 897N203029A FDC NX600 Instruction Manual 2022.01 6-17 6 6 Maintenance 6.4 Replacing Recording Paper When red lines appear along the sides of the recording paper, the printer is nearly out of paper, so replace the recording paper roll with a new one. NOTE Use the recording paper specified for the FDC NX600. 1. Press the section indicated in the illustration (1) to release the lock, and then open the cover (2). 1 2 Press the concave section. 2. Remove the remaining recording paper. 6 3. Load the new recording paper. NOTE If the paper replacement was performed during a measurement, the measurement results might be printed out. NOTE Set the paper roll so the paper is pulled from under the paper roll as shown below. OK Incorrect NOTE If the paper is loaded in reverse, no text is printed on the thermal paper. 4. Pull the top end of the recording paper out slightly, and then close the cover. NOTE When setting a new recording paper, pull the edge of paper out to where the tape for fixing edge is affixed. Failure to observe this precaution may cause printing malfunction. NOTE Close the cover securely until it clicks. Otherwise, the paper feed may fail and the test results will not be printed out. 6-18 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.5 Replacing and Cleaning the Light Source Lamp The average life of the lamp is about 1,000 hours. Replace the lamp if a lamp replacement error occurs or the cumulative lighting time of the lamp approaches 1,000 hours. CAUTION: The light source lamp gets very hot. Before cleaning the light source unit or replacing the light source lamp, turn off the power, wait at least 5 minutes, and then check the lamp has cooled. 1. Pull out the disposal box, and then remove the front bottom cover. 6 Front bottom cover 2. Loosen the thumbscrew as shown in the left figure and remove the lamp unit. 3. Unplug the connector of the light source lamp and remove the lamp unit. 1 Thumbscrew 3 2 Lamp unit Connector 897N203029A FDC NX600 Instruction Manual 2022.01 6-19 6 Maintenance 4. Wipe the light source unit lens using a dry cotton swab. Lens 5. Turn the thumbscrew of the lamp unit to the left, remove the lamp, and then replace it with a new one. Lamp base IMPORTANT: Tighten the thumbscrews securely. Otherwise, the measurement results may be negatively affected. 6 Thumbscrew IMPORTANT: NOTE When you replace the lamp, hold it by its base. Do not touch the glass surface with bare hands. Secure so that metal hooks hold from above the lamp housing. 6-20 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance Thumbscrew B 2 3 1 6. Insert the connector of the light source lamp. 7. Insert the lamp unit into the analyzer with thumbscrew B facing up. Tighten thumbscrew A securely. IMPORTANT: Tighten the thumbscrew securely. Otherwise, the measurement results may be negatively affected. Thumbscrew A Connector 8. Attach the front bottom cover, load the disposal box, and then switch on the power of the FDC NX600. NOTE Put the front bottom cover back horizontally as it was. IMPORTANT: Be sure to load the disposal box. If the analyzer is used without the box, the measurement results may be negatively affected. 9. After starting the analyzer, go to [Lamp management] in Normal mode, and then reset [Lamp ON time]. è Refer to Section 9.2.3 897N203029A FDC NX600 Instruction Manual 2022.01 6-21 6 6 Maintenance 6.6 Maintenance of Sampler O-ring The sampler nozzle O-ring wears over time. Perform weekly cleaning, monthly inspection, and annual replacement. WARNING: When inspecting and replacing O-rings, always wear gloves, a lab coat, and safety goggles. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: If you do not inspect and replace the sampler O-ring, the sample application volume will be inaccurate, which may negatively affect the measurement values. 6.6.1 Cleaning the Sampler O-ring Clean the sampler O-ring once a week as follows. 1. Prepare a dry / damp gauze or cloth (eyeglass wipes, lint-free cloth, etc.). 6 2. Switch off the analyzer before cleaning. 3. Rotate the sampler part toward you. 6-22 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 4. While holding the arm with one hand, wipe the O-ring while rotating the damp cloth 3 to 4 turns around the circumference. Next, wipe down from the top to cover the entire circumference. Similarly, wipe the other O-ring around the circumference and down from the top. 5. Use a dry cloth to lightly wipe the moisture from the O-ring portion. Remove any foreign objects remaining on the O-ring. 6. Perform a measurement and confirm that no errors occur. IMPORTANT: Do not store the AUTO TIPS with the storage box lid open. If AUTO TIPS contaminated with dust or dirt are mounted to the nozzles, the dirt or dust may adhere to the O-rings, causing dispensing/ dilution clogging errors or leak check failure. IMPORTANT: If dirt is noticeable or an error occurs, shorten the cleaning cycle (once a week or more). 7. If an error occurs, inspect the sampler O-ring (Refer to Section 6.6.2). 897N203029A FDC NX600 Instruction Manual 2022.01 6-23 6 6 Maintenance 6.6.2 Inspecting the Sampler O-ring 1. Tap the button. 2. Tap the [Normal mode] button. 6 3. Tap [Sampler leak check]. 6-24 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 4. Follow the on-screen instructions to prepare for the leak check. 5. Tap the nozzle to be checked to select it, and then run the leak check. Once check ends, the Check results screen appears. 6 CAUTION: Keep your fingers clear as the sampler unit and consumable disk move. 6. Check the results of the [Sampler leak check]. If the check results are [OK], it is not necessary to replace the O-ring. NOTE The check results can be printed by button. tapping the 897N203029A FDC NX600 Instruction Manual 2022.01 6-25 6 Maintenance If the check results are [NG], tap the [Replace O-ring] button, and then replace the O-ring according to the following procedure. NOTE The check results can be printed by button. tapping the 7. Once the check ends, open the sampler cover, and then remove the sampler leak check tool mounted on the sampler nozzle by hand. 6 6-26 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.6.3 Exchanging the Sampler O-ring The sampler nozzle O-ring wears over time. Replace it once a year. WARNING: Process used O-rings as infectious waste. IMPORTANT: If you do not replace the O-ring periodically, the sample application volume will be inaccurate, which may negatively affect the measurement values. IMPORTANT: Be sure to use the O-ring designed for the FDC NX600. Otherwise, the sample application volume will be inaccurate, and the measurement results may be negatively affected. 1. Remove the sample rack and slide cartridge. 2. Tap the button. 6 3. Tap the [Normal mode] button. 897N203029A FDC NX600 Instruction Manual 2022.01 6-27 6 Maintenance 4. Tap [Replace O-ring]. 5. Tap the [START] button. The sampler nozzles turn to the replacement position. 6 CAUTION: Keep your fingers away as the sampler unit moves. 6. Follow the on-screen instructions to replace the O-ring. If this button is tapped, the following explanation appears. 6-28 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 7. Tap [Leak check] to check for leaks. è Refer to Section 6.6.2 6 897N203029A FDC NX600 Instruction Manual 2022.01 6-29 6 Maintenance 6.7 Cleaning the Slide Reading Section When a slide-detection error has occurred, dust and other soiling may be the cause. Clean the slide reading section to avoid these errors. WARNING: Always wear gloves, a lab coat, and safety glasses while working. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Dispose of contaminants used for cleaning as infectious waste. 1. Switch off the analyzer power. 2. Open the sampler cover and panel unit. 3. Remove the slide cartridge. 6 RE 4. Using your finger, move the transfer bar under the right cover. RE 5. Air blow from the gap where the slide cartridge is placed, as shown in the left figure. RE 6-30 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.8 Cleaning and Replacing the Reference Fluid Cap Packing Clean the reference fluid cap packing each time the reference fluid is replaced. Replace the reference fluid cap packing once every three months. IMPORTANT: If you do not clean and replace the reference fluid cap packing, the reference fluid will be concentrated, which may negatively affect the measurement values. 1. If reference fluid is loaded, remove the reference fluid tube and discard it. 2. Remove the reference fluid cap packing as shown in the illustration on the left. 6 Reference fluid cap packing Cotton swab 3. Wipe off the reference fluid cap packing using a cotton swab or gauze moistened with lukewarm or cold water as shown in the illustration on the left, and then allow it to dry. NOTE Replace it with a new cap if stains cannot be wiped off. Reference fluid cap packing 897N203029A FDC NX600 Instruction Manual 2022.01 6-31 6 Maintenance 4. Set the reference fluid cap packing properly. NOTE Do not touch the bottom side (reference fluid side) of the reference fluid cap packing. Reference fluid cap packing 6 6-32 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.9 PF Unit Maintenance WARNING: During maintenance, always wear gloves, a lab coat, and safety glasses. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. WARNING: Dispose of contaminants used for cleaning as infectious waste. 6.9.1 Cleaning the PF Unit (Sensor, Reference Plate, and Suction Pad) Clean when errors occurred during PF filtering. 1. Tap the button. 6 2. Tap the [Normal mode] button. 897N203029A FDC NX600 Instruction Manual 2022.01 6-33 6 Maintenance 3. Tap [PF maintenance]. 4. Tap the [Cleaning] button. 6 5. Remove the sample rack, mixing cup, and tip rack. 6. Tap the [OK] button. CAUTION: Keep your fingers clear as the PF part moves. 6-34 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 7. Clean according to the on-screen instructions. 8. Tap the [OK] button to end cleaning. Surface of PF sensor (glass surface) 6 PF Suction Pad (black) PF reference plate 9. Run the leak check. è Refer to Section 6.9.2 NOTE If the check was [NG], exchange the PF suction pad. è Refer to Section 6.9.3 897N203029A FDC NX600 Instruction Manual 2022.01 6-35 6 Maintenance 6.9.2 Inspecting the PF Suction Pad The PF suction pad must be inspected monthly and replaced every two years. 1. Touch the [PF Maintenance] button as in the previous section. 2. Tap the [Leak check] button. 6 6-36 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 3. Follow the on-screen instructions to load the PF sample rack and PF leak check tool to the sample position, and then tap the [START] button to run the leak check. If check ends, the Check results screen appears. CAUTION: Keep your fingers clear as the PF part moves. 4. Check the results of the [Leak check]. If the check results are [OK], it is not necessary to clean or replace the pad. NOTE The check results can be printed by tapbutton. ping the If the check results are [NG], tap the [Cleaning] button, clean the PF unit, and then run the leak test again. è Refer to Section 6.9.1 If this does not resolve the problem, tap the [Procedure] button, and then check the assemblies of the PF pump tube or replace the tube, referring to the displayed instructions. NOTE If the [Result details] button is tapped, the Advanced leak check results display screen appears. NOTE The check results can be printed by tapbutton. ping the 5. Once the check ends, remove the PF sample rack and PF leak check tool. 897N203029A FDC NX600 Instruction Manual 2022.01 6-37 6 6 Maintenance 6.9.3 Replacing the PF Suction Pad Replace a PF suction pad once every 2 years. WARNING: Process used PF Suction Pad as infectious waste in compliance with any applicable regulations in your country. 1. Touch the [PF Maintenance] button as in the previous section. 6 2. Tap the [Suction pad] button. The PF sampler rotates forward and stops. 6-38 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 3. Remove the sample rack, mixing cups, and tip rack. RE Sample rack Tip rack 6 Mixing cups 4. Tap the power. 897N203029A FDC NX600 Instruction Manual 2022.01 button to switch off the 6-39 6 Maintenance 5. Turn the PF Suction Pad in the direction shown and remove it while loosening. 6. Install the new PF Suction Pad by turning it in the opposite direction. NOTE Be sure to tighten the PF Suction Pad securely. 7. Switch on the analyzer. 8. Run the leak check. RE If the check results are [NG], perform the replacement again. è Refer to Section 6.9.2 6 6-40 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 6.9.4 Replacing the PF Pump Tube WARNING: Before the following procedure, always switch off the power of the analyzer, and unplug the power cord. Replace the PF pump tube once every 2 years. Replacement is also necessary if the PF has not been used for one month or longer. 1. Remove the disposal box, and remove the front bottom cover. 6 Front bottom cover Cover (B) Holder (C) Screw (A) Tube (D) 2. Loosen the screw (A) and remove the cover (B). 3. Slide the holder (C) to the left and remove the tube (D) from the rotor. 897N203029A FDC NX600 Instruction Manual 2022.01 6-41 6 Maintenance 4. While pushing the head of joint (G), pull the joint (G) to the right to remove the tube (D). PF pump Joint (G) Red tube (D) Pull off joint (G) to the right Joint (G) while pushing this part to the right. Holder (C) 5. Slide the holder (C) to the left and load the new tube (H) on to the rotor. Tube (H) NOTE As shown in the diagram to the left, align the base of the joint (E) with the cover hole (F). NOTE Set so that the two front marks face the front. 6 6. Reattach the cover (B) using the screw (A). Cover hole (F) Joint (E) Front Marks Joint (E) Cover hole (F) When attaching the cover (B), make sure that the hooks are hooked and interlocked. NOTE If attachment is incorrect, a “W0134 PF sampling error” may occur. Red tube (H) Joint (G) 7. Insert the tip of the red tube (H) in the joint (G) until the line. Line 6-42 897N203029A FDC NX600 Instruction Manual 2022.01 6 Maintenance 8. Attach the front bottom cover. NOTE Put the front bottom cover back horizontally as it was. 9. Attach the disposal box. 10. After the replacement, run the leak check. If the check results are [NG], perform the replacement again. è Refer to Section 6.9.2 6 897N203029A FDC NX600 Instruction Manual 2022.01 6-43 7 Quality Control IMPORTANT: To maintain the accuracy of your test results, it is recommended to perform quality control on routine bases by use of control fluids. Perform quality control in compliance with local regulation and procedure which you have to obey. Also, it is also recommended to do after major repairs are performed on the analyzer. 7.1 Control Fluids Use the controls which are prepared for FUJI DRI-CHEM or which are recommended to use. NOTE Follow the descriptions in the “Instructions for Use” of control fluids. 7.2 Measurement with Control Fluids Measure with control fluids by following the same procedure as for measurement of patient samples. Load the control fluid in the sample loading position of the sample rack, load a slide, and then measure. 7.3 Monitoring Results Configure the target values and acceptable ranges for measurements from control fluid results, and monitor for changes over time. 7.4 Control Mode You can measure control liquid by using [Control] of the Function menu to configure the entered correlation coefficients “a” and “b” to a = 1 and b = 0 temporarily. This state is called Control mode. è Refer to Section 9.1 After using this mode, always cancel control mode by tapping [HOME] on the LCD. 7.5 Troubleshooting Quality Control Results If there is a discrepancy in the control fluid measurement results, confirm the following points, and then measure the control fluid again. • Make sure that the periodic cleaning and replacement of consumables listed in this manual are being performed. • Make sure the handling methods, storage conditions, usage limitations, and other details listed in the “Instructions for Use” of slides have been observed. If the control results are still a problem, please contact your dealer. 897N203029A FDC NX600 Instruction Manual 2022.01 7-1 7 7 Quality Control 7.6 Control Chart 1. Tap the button. 2. Tap the [Control] button. 7 3. The measurement-preparation screen of control mode appears. 4. Tap the 7-2 (graph display) button. 897N203029A FDC NX600 Instruction Manual 2022.01 7 Quality Control 5. Tap the button of the measurement item to be viewed. 6. The control chart appears. NOTE If the control liquid button is tapped, you can change the control liquid to be displayed. NOTE Measurement results of outside of measurement range (@), disposal box open (*), and expired slide (#) are not reflected in the control chart. NOTE If the measurement date is touched and selected, and the show/hide button is touched, the plot of these measurement results can be hidden (or shown). Show/hide button Acceptable range lower limit Acceptable range upper limit Measurement result average Outside of display area Control lot was changed Control lot was changed Outside of acceptable range 897N203029A FDC NX600 Instruction Manual 2022.01 7-3 7 7 Quality Control 7. If a measurement date is selected and then the [Detail] button is tapped, the detail screen appears. < Display of measurement results > Green: when within the acceptable range of control Red: when outside of the acceptable range of control 7 Acceptable range lower limit Acceptable range upper limit NOTE The acceptable range of control displays the value entered for Range setting for control (Refer to Section 9.3.17). NOTE In case of ALPi and LDHi slides, the display of “Test name” will not follow the “Test Item Name Setting”, but “ALPi-PS” or “LDHi-PS”. But, the test name of each measurement result will follow the “Test Item Name Setting”. è Refer to Section 9.3.20. 7-4 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting WARNING: When you troubleshoot, always wear gloves, a lab coat, and safety goggles. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: If an analyzer malfunction (error) appears before or during measurement or warnings appear in the measurement results, the measurement results may be inaccurate. Refer to the related troubleshooting pages and rerun the measurement. 8.1 Error Display If an error occurs, an error message appears on the LCD. < Example of display > Error code Content of error message Additional information NOTE Some error codes may show additional information. (Example) 1050 (GLU-P slide code) 120001 (Lot number) *Refer to Section 8.3 for slide codes. NOTE The error messages listed in this document and the descriptions displayed on the LCD may differ. 897N203029A FDC NX600 Instruction Manual 2022.01 8-1 8 8 Troubleshooting 8.1.1 Error Log You can view the history of errors that occurred by using [Error log] in the function menu. 1. Tap the button. 2. Tap the [Error log] button. An error log list is displayed. 8 3. Tap an error. The error details are displayed. 8-2 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-3 8 Troubleshooting 8.1.2 Error Code and Reference Section Table Error code 8 8-4 Error description Reference section Reference page E0021 The disposal box was open during measurement 8.2.15 8-42 E0035 Photometric control system error 8.2.8 8-24 E0050 Light source lamp is burned out 8.2.5 (2) 8-20 E0080 Tip detection sensor error 8.2.15 8-42 E0082-E0085 Tip mounting error 8.2.4 (1) 8-11 E0086, E0087 Tip disposal error 8.2.4 (2) 8-12 E0111, E0112 Clogging error during sampling of sample 8.2.4 (5) 8-14 E0121, E0122 Clogging error during sampling of diluent 8.2.4 (8) 8-17 E0124 Clogging error during sampling of diluted sample 8.2.4 (5) 8-14 E0125 Clogging error during sampling of diluted sample 8.2.4 (5) 8-14 E0127, E0128 Clogging error during sampling of reference fluid 8.2.4 (8) 8-17 E0132, E0133, E0150 PF errors 8.2.12 8-36 E0154 Mixing cup error 8.2.4 (3) 8-13 E0160 Incorrect diluent type error 8.2.10 8-32 E0200-E0203, E0210-E0213, E0220-E0227 Slide transfer error 8.2.6 8-21 E0300, E0301, E0302 Reference white plate measurement error during initial operation 8.2.8 8-24 E0509, E0510 Signal system error inside circuit board 8.2.8 8-24 E0511 Light intensity error 8.2.5 (1) 8-19 E0515, E0516, E0517 Signal system error inside circuit board 8.2.8 8-24 E0530, E0531 ISE measurement unit (self-test) error 8.2.11 (1) 8-33 E0532, E0533, E0537, E0538 Signal system error inside circuit board 8.2.8 8-24 E0534, E0535, E0536, E0539, E0540 Signal system error inside circuit board 8.2.8 8-24 E0565 ISE temperature control error 8.2.7 8-22 E0600-E0604 Incubator motor error 8.2.6 8-21 E0700-E0703 ISE measurement unit motor error 8.2.6 8-21 E0900-E0903 Interference filter motor error 8.2.5 (3) 8-20 E1000-E1003 Sampler vertical operation error 8.2.4 (9) 8-18 E1100-E1103 Sampler rotation error 8.2.4 (9) 8-18 E1200-E1203 Syringe operation error 8.2.4 (9) 8-18 E1300-E1304 Errors related to PF 8.2.12 8-36 E1400-E1403 Sample application unit feed error 8.2.6 8-21 E1500, E1501, E1505 Temperature control error 8.2.7 8-22 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting Error code Error description Reference section Reference page E1502, E1503, E1504 Temperature control error 8.2.7 8-23 E1510, E1511, E1512, E1514, E1515 ISE temperature control error 8.2.7 8-22 E1520, E1521 Installation environment temperature error 8.2.7 8-22 E1530, E1531, E1532 Signal system error inside circuit board 8.2.8 8-25 E1600-E1603 Consumable disk operation error 8.2.6 8-21 E1750-E1755, E1757 Slide reading control system error 8.2.8 8-25 E3900 Signal system error inside circuit board 8.2.8 8-25 E4200, E4201 Calibration error 8.2.14 (3), (4) 8-39 E4202, E4203 Calibration error 8.2.14 (5) 8-40 E4204 Calibration error 8.2.14 (6) 8-41 E4205 Calibration error 8.2.14 (5) 8-40 E4401, E4403 Signal system error inside circuit board 8.2.8 8-25 E4410 Measurement result reference error 8.2.8 8-25 E4411 Control chart reference error 8.2.8 8-25 E4454, E4475, E4478-E4481, E4490, E4600E4613 Signal system error inside circuit board 8.2.8 8-26 E5101, E5102, E5103 Printer error 8.2.2 (3) 8-10 E5500 Signal system error inside circuit board 8.2.8 8-28 E5510 Slide position offset during measurement 8.2.5 (4) 8-20 E5901 Signal system error inside circuit board 8.2.8 8-28 E6200, E6201, E6210 Communication error 8.2.13 (1) 8-37 897N203029A FDC NX600 Instruction Manual 2022.01 8 8-5 8 Troubleshooting Error code 8 8-6 Error description Reference section Reference page W0020 The disposal box is open 8.2.15 8-42 W0022 The disposal box is full 8.2.15 8-42 W0030 Insufficient light intensity 8.2.5 (1) 8-19 W0036 Reference black plate soiling 8.2.5 (1) 8-19 W0040, W0041 The sample type configuration and sample type of the slide do not match 8.2.10 8-31 W0042, W0043 Sample rack configuration error 8.2.10 8-31 W0060 Insufficient sample or calibrator volume 8.2.4 (4) 8-13 W0070 Sampler rotation error 8.2.4 (9) 8-18 W0088, W0089, W0096 Sample tip detection error 8.2.4 (1) 8-11 W0090 The ISE slide is loaded in reverse 8.2.10 8-31 W0091, W0092 ISE measurement error 8.2.11 (3) 8-35 W0120 Sample surface cannot be detected 8.2.4 (4) 8-13 W0123, W0129 Liquid volume error 8.2.4 (7) 8-16 W0134, W0138 PF errors 8.2.12 8-36 W0137 PF card has not been read 8.2.9 (3) 8-29 W0140 Liquid volume error 8.2.4 (4) 8-13 W0141-W0144 Liquid volume error 8.2.4 (7) 8-16 W0151, W0152 Mixing cup error 8.2.4 (3) 8-13 W0170 QC card has not been read 8.2.9 (2) 8-29 W0173 The slide information has not been read 8.2.3 8-11 W0175 No slides have been set 8.2.10 8-31 W0180 Calibration has not been performed with an immuno-chemical test 8.2.14 (1) 8-39 W0190, W0191 Calibration error 8.2.14 (2) 8-39 W0214 Slide transfer error 8.2.6 8-21 W0500-W0503 Malfunction of the internal fan 8.2.15 8-42 W0704 ISE measurement unit motor error 8.2.6 8-21 W0904 Interference filter motor error 8.2.5 (3) 8-20 W1004 Sampler vertical operation error 8.2.4 (9) 8-18 W1104 Sampler rotation error 8.2.4 (9) 8-18 W1204 Syringe operation error 8.2.4 (9) 8-18 W1308 PF errors 8.2.12 8-36 W1404 Sample application unit feed error 8.2.6 8-21 W1604 Consumable disk operation error 8.2.6 8-21 W1700 Malfunction of the QC card reader 8.2.9 (5) 8-30 W1703 QC card read ERR 8.2.9 (5) 8-30 W4010 QC type has not been read 8.2.9 (2) 8-29 W4011 Panel card error 8.2.9 (5) 8-30 W4012 DI card error 8.2.9 (4) 8-30 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting Error code Error description Reference section Reference page 8.2.9 (5) 8-30 W4013, W4014, W4015 QC card error W4440 Setting value reset warning 8.2.15 8-43 W4460-W4475, W4477, W4483 Signal system error inside circuit board 8.2.8 8-26 W4500, W4501, W4502 Serial communication error 8.2.13 (1) 8-37 W4510 LAN communication error 8.2.13 (1) 8-37 W4511 Type 4 LAN communication error 8.2.13 (1) 8-37 W4520 Error related to work list 8.2.13 (1) 8-37 W4720 Clock time error 8.2.8 8-27 W4721, W4733 Signal system error inside circuit board 8.2.8 8-28 W5000, W5020 Signal system error inside circuit board 8.2.8 8-28 W5002-W5019 USB memory read error 8.2.8 8-28 W5050 USB memory has not been inserted 8.2.8 8-28 W5100 Printer error 8.2.2 (3) 8-10 W5110 No recording paper 8.2.2 (3) 8-10 W5330-W5332 Sample barcode reader error 8.2.13 (2) 8-38 W5400 Expired slide 8.2.10 8-32 W66xx Input value waring 8.2.15 8-43 I5920 Reboot warning 8.2.15 8-42 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-7 8 Troubleshooting 8.1.3 Printout Reference Table Printout H L > < @ Meaning and conditions Measurement value is higher than the upper limit of the preset reference interval. Measurement value is lower than the lower limit of the preset reference interval. Measurement value is higher the upper limit of the measurement range. Measurement value is lower than the lower limit of the measurement range. The measurement values are reference values outside the measurement range. Troubleshooting - - IMPORTANT: Measurement values might be incorrect. + OR - OR +, -, $ *, ? E & 8 8-8 Measurement value of the ISE test is higher than the upper limit of the measurement range. Measurement value of the ISE test is lower than the lower limit of the measurement range. Temperature control error during measurement. Light intensity is fluctuating during measurement. Unable to calculate the measurement result because measurement was interrupted due to a transfer problem or other error. The measurement results are printed as [****]. Abnormally high measurement value. The glucose value of the sample is high in a CRP measurement. In the case of a Ca-P slide, the sample has not been applied. Refer to Section 8.2.11 Refer to Section 8.2.11 Refer to Section 8.2.7 Refer to Section 8.2.5 Check whether other errors have occurred, and then perform the necessary troubleshooting. Rerun the measurement. Perform a dilution measurement by following the “Instructions for Use” of the slides. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting # The valid term of the slide has expired. For immuno-chemical tests, the validity term of the slide or the validity term of calibration has expired. IMPORTANT: The measurement value might be incorrect. IMPORTANT: Expired slides cannot be used. Always use valid slides. IMPORTANT: Use slides within their validity terms. Analyzer will not show # whose means that ISE slides have been expired if the installed software is version 1.0. Otherwise, with the software version 1.1 or later is installed, analyzer will show # for the ISE slides that expire 2023 or later or of lot number 130000 to 139999. ¥ NOTE Depending on the device settings, there may be cases in which # is not displayed, and an expired slide cannot be measured. Problem with sample application on slide (not applied) Rerun the measurement. (Supplement) Measurement range The measurement range (determination range) and the reference interval differ by measurement item. Their relation is shown below. For details, refer to the “Instructions for Use” of the slides. NOTE You can set the values of the reference interval by using [Reference interval range setting] in Mode function for administrator. (Refer to Section 9.3.6) High Measured value Reference interval Measurement range (Determination range) Low 897N203029A FDC NX600 Instruction Manual 2022.01 8-9 8 8 Troubleshooting 8.2 Troubleshooting 8.2.1 Startup Errors (1) The analyzer does not start after the power button is pressed. Check that the power cable is properly connected. And, turn the power switch OFF and ON again, and press the power button. If the analyzer still does not start, please contact your dealer for assistance. 8.2.2 Printer or LCD Trouble (1) Printed characters are too faint or dark Please contact your dealer for assistance. (2) White lines appear on printed characters or certain parts of characters are missing. The printer head has malfunctioned. Please contact your dealer for assistance. (3) Printer errors Error code 8 Error message Error condition Troubleshooting W5110 Printer error No paper. Set paper. No recording paper. Load recording paper onto the printer. è Refer to Section 6.4 W5100, E5101, E5102, E5103 Printer error Contact your dealer in case the ERR repeats. The printer is malfunctioning. Please contact your dealer for assistance. (4) LCD is too faint or dark Adjust the brightness using [Brightness • Volume] in Normal mode. è Refer to Section 9.2.2 8-10 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.3 Slide Reading Errors Error code W0173 Error message Error condition Slide identification ERR No slide information. Check the slide. The slide information on backside of the slide could not be read correctly. Troubleshooting Clean the slide reader, and then rerun the measurement. è Refer to Section 6.7 NOTE If the error occurs again, please contact your dealer for assistance. 8.2.4 Sampler Errors (1) When tip cannot be mounted Error code E0082, W0088 E0083, W0089 Error message Error condition Sample tip detection No tip can be detected on the sample nozzle ERR after mounting. No tip detected. Set tips and rerun the tests from the beginning. Reference fluid tip detection ERR No tip detected. Set tips and rerun the tests from the beginning. No tip has been detected on the reference fluid nozzle after mounting. Troubleshooting The tip is not loaded correctly. Load a tip correctly, and then rerun the test from the beginning. è Refer to Section 5.2 NOTE Set tips into all holes of the tip rack after this error has occurred. 8 E0084 If the error occurs again, please contact Sample tip clogging A clog was detected during mounting of the your dealer for assistance. ERR sample nozzle tip. Rerun the tests from the beginning. Contact your dealer in case the ERR repeats. E0085 Reference fluid tip clogging ERR Rerun the tests from the beginning. Contact your dealer in case the ERR repeats. A clog was detected during mounting of the reference fluid nozzle tip. W0096 No tip WARNING Tip is insufficient. Set tips and rerun the test from the beginning. The number of tips needed for measurement is insufficient. 897N203029A FDC NX600 Instruction Manual 2022.01 Load tips in all holes of the tip rack. 8-11 8 Troubleshooting (2) When tip cannot be ejected Error code Error message Error condition E0086 Sample tip eject ERR Tips cannot be discarded. After analyzer shutdown, please check the disposal box. Turn the power ON. E0087 Reference fluid tip eject ERR After analyzer shutdown, please check the disposal box. Turn the power ON. Troubleshooting After analyzer shutdown, empty disposal box, and then turn the power ON. 8 8-12 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting (3) When there are insufficient number of or no mixing cups Error code Error message Error condition Troubleshooting W0151 There are not enough Lack of mixing cup mixing cups. Lack of mixing cups. Set mixing cup and rerun the tests from the beginning. W0152 No mixing cup Set mixing cup and rerun the tests from the beginning. E0154 Mixing cup detection Error of the mixing cup If the error still occurs after switching sensor ERR detection sensor. the analyzer off and on, please contact your dealer for assistance. Contact your dealer. After you replace the mixing cups, rerun from the measurement request. è Refer to Section 5.3 (4) During a sample volume error Error code Error message Error condition Troubleshooting Either the sample is not loaded or the sample volume is insufficient. Make sure that the sample volume in the tube is within the measurable range and an appropriate sample tube is used in the sample rack, and then rerun from the measurement request. W0120 Sample shortage Sample is insufficient. Check the sample. W0060 Sample shortage There is not enough Check the sample vol- sample volume. ume. Confirm that a sample rack matching the sample tubes is used. W0140 S u r f a c e d e t e c t i o n There is too much ERR (sample) sample volume. Sample volume is too much. Check the sample volume. 897N203029A FDC NX600 Instruction Manual 2022.01 NOTE If the sample volume is low, add sample to the tube or perform manual sample application. 8-13 8 8 Troubleshooting (5) During a sample aspiration or application error Error code E0111 Error message Error condition Aspiration clogging Clogging was detected ERR (sample) during sample aspiraClogging was detected tion. during sampling. Check the sample. E0124 Aspiration clogging ERR (diluted sample) Clogging was detected during sampling. Check the sample and diluent. E0112 S p o t t i n g c l o g g i n g A clog was detected ERR (sample) during sample applicaClogging was detected tion (dispensing). during sampling. Check the sample. E0125 S p o t t i n g c l o g g i n g A clog was detected ERR (diluted sample) during sample applicaClogging was detected tion (dispensing). during sampling. Check the sample and diluent. 8 8-14 Troubleshooting The sample may contain fibrin deposits. After checking the sample, rerun from the measurement request. NOTE If the error occurs again, there might be a piping system problem. Please contact your dealer for assistance. The sample may contain fibrin deposits. After checking the sample, rerun from the measurement request. NOTE If the error occurs again, there might be a piping system problem. Please contact your dealer for assistance. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting (6) When application is poor Error code Error message Sample application problem Error condition Troubleshooting Because the sample is Check the following points. not applied to the slide, (a) Make sure that FUJI DRI-CHEM the error mark [<] or [@] AUTO TIPS are being used. appears continuously. (b) Make sure that tips are not being reused. IMPORTANT: (c) Make sure that there are no bubbles on the sample surface in the Measurement values sample tube. with an [@] mark (d) Make sure that a sample rack that might be incorrect. matches the blood collection tube Run the measureor FUJI tube is being used. ment again. (e) Inspect the sampler O-ring. (Refer to Section 6.6.2) IMPORTANT: FUJI DRI-CHEM AUTO TIPS cannot be reused. NOTE If errors still occur often, please contact your dealer for assistance. 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-15 8 Troubleshooting (7) During a diluent/reference fluid volume error Error code Error message Error condition W0141 Surface detection ERR There is too much dilu(diluent) ent volume. Diluent volume is too much. Check the diluent volume. W0142 Surface detection ERR There is too much ref(reference fluid) erence fluid volume. Reference fluid is too much. Check the reference fluid volume. W0123 Diluent is insufficient. There is not enough Diluent is insufficient. diluent. Check the diluent. W0143 Diluent is insufficient. Diluent is insufficient. Check the diluent volume. W0129 Insufficient reference There is not enough fluid reference fluid. Reference fluid is insufficient. Check the reference fluid. W0144 Insufficient reference fluid Reference fluid is insufficient. Check the reference fluid volume. 8 8-16 Troubleshooting Follow the instructions of the error code to replace the diluent or reference fluid, and then rerun from measurement request. • Diluent è Refer to Section 5.4 • Reference fluid è Refer to Section 5.5 NOTE Replace the diluent / reference fluid container with a new one. Do not add diluent or reference fluid. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting (8) During a diluent / reference fluid error Error code E0121 E0122 Error message Error condition Troubleshooting Aspiration clogging Clogging was detected ERR (diluent) during diluent aspiraClogging was detected tion. during sampling. Check the diluent. Be sure that FUJI tubes configured in Tube setting for diluent and reference fluid (è Refer to Section 9.3.15) are used. If there was a loading problem, replace the diluent and reference fluid, and then rerun from the measurement Spotting clogging ERR Clogging was detected request. (diluent) during diluent applica- NOTE If the error occurs again, there Clogging was detected tion. might be a piping system probduring sampling. lem. Please contact your dealer Check the diluent. for assistance. E0127 Aspiration clogging Clogging was detected ERR (reference fluid) during reference fluid Clogging was detected aspiration. during sampling. Check the reference fluid. E0128 Spotting clogging ERR Clogging was detected (reference fluid) during reference fluid Clogging was detected application. during sampling. Check the reference fluid. 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-17 8 Troubleshooting (9) During a sampler drive system error Error code W0070 E1000, E1001, E1003, W1004 8 Error message Error condition S p o t t i n g p o s i t i o n Rotation error of the Turn the analyzer off after the measERR sampler has occurred. urement, and make sure that foreign matters such as tips and slides are not Sampler rotation error. stuck or left inside of the sampler. S a m p l e r u p / d o w n Up-and-down moveNOTE If the error still occurs after motor ERR ment error of the motor switching the analyzer off and Problem moving the on, please contact your dealer sampler up and down. for assistance. Perform inspection. Contact your dealer. E1002 Sampler up/down motor operation ERR Problem moving the sampler up and down. Perform inspection. Contact your dealer. E1100E1103, W1104 Sampler movement Rotation error of the motor ERR sampler Problem with sampler movement. Perform inspection. Contact your dealer. E1200, E1201, E1203, W1204 Syringe motor ERR Problem with the syringe. Contact your dealer. E1202 Syringe motor operation ERR Problem with the syringe. Contact your dealer. 8-18 Troubleshooting Movement error of the syringe motor 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.5 Photometric System Errors IMPORTANT: If a photometric system error has occurred, perform each troubleshooting step listed below and rerun the measurement. (1) During a fluctuation of light source intensity Error code W0036 W0030 Error message Error condition Soiled black plate The reference plate is WARNING dirty. Reference black plate soiled. Clean the black plate. Troubleshooting Switch off the analyzer and clean each part listed below. • Clean the light source unit lens è Refer to Section 6.5 • Clean the reference plate Lamp replacement The light intensity of WARNING the light source lamp is è Refer to Section 6.3 insufficient. Low light intensity. • Clean the photometer Clean photometer. è Refer to Section 6.3 After cleaning, switch on the power. If the errors still occur after cleaning, replace the light source lamp. è Refer to Section 6.5 NOTE If the error occurs again, please contact your dealer for assistance. E0511 Reference white plate The light intensity is CAUTION: ERR changing during measurement. Contact your dealer. The light source lamp gets very hot. Before cleaning the light source unit or replacing the light source lamp, turn off the power, wait at least 5 minutes, and then check the lamp has cooled. Error marks “ * ” or “ ? ” are printed on the results 897N203029A FDC NX600 Instruction Manual 2022.01 8-19 8 8 Troubleshooting (2) When lamp is burned out Error code E0050 Error message Lamp replacement Low light intensity. Replace the lamp. Error condition The light source lamp is burned out. Troubleshooting Switch off the power and replace the light source lamp. Switch the analyzer on again after replacement. è Section 6.5 CAUTION: The light source lamp gets very hot. Before cleaning the light source unit or replacing the light source lamp, turn off the power, wait at least 5 minutes, and then check the lamp has cooled. (3) During an interference filter motor error Error code Error message Error condition Troubleshooting E0900, E0901, E0903, W0904 Filter motor ERR Movement error of the Switch the analyzer off and on. Problem with the inter- filter motor NOTE When the error is not resolved, ference filter. please contact your dealer for Contact your dealer. assistance. E0902 Filter motor operation ERR Problem with the interference filter. Contact your dealer. 8 (4) Other errors Error code E5510 8-20 Error message Slide slip ERR Contact your dealer. Error condition Troubleshooting The slide stop position Contact your dealer and have the incushifted during measbator pressure plate inspected. urement. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.6 Transfer Errors Error code Error message E0200E0203, E0220E0227 Feed motor ERR Problem with the slide transfer system. Perform inspection. Contact your dealer. E0210E0213, W0214 Eject motor ERR Problem with the slide transfer system. Perform inspection. Contact your dealer. Error condition Troubleshooting Movement error of the The sample application unit or the incuslide supply motor bator may be dirty or the disposal box may be full of slides and tips. Switch off the analyzer, and perform the following. • Cleaning the disposal box • Cleaning the incubator è Refer to Section 6.3 • Cleaning the sample application unit è Refer to Section 6.3 E0600E0604 Incubator motor ERR Movement error of the Problem with the incu- incubator motor of the incubator bator. Perform inspection. Contact your dealer. E0700, E0701, E0703, W0704 Probe motor ERR Problem with ISE. Perform inspection. Contact your dealer. E0702 Probe motor operation ERR Problem with ISE. Perform inspection. Contact your dealer. NOTE If the error occurs again, please contact your dealer for assistance. Movement error of the probe motor of the ISE unit 8 E1400E1403, W1404 Transfer change motor ERR Problem with the slide transfer system. Contact your dealer. E1600E1603, W1604 C o n s u m a b l e d i s k The disk motor movemotor ERR ment error in consumProblem with the con- able disk motor sumable disk. Contact your dealer. Movement error of the ISE transfer motor of the sample application unit 897N203029A FDC NX600 Instruction Manual 2022.01 Turn the analyzer off after the measurement, and make sure that foreign matters such as tips and slides are not stuck or left around the consumable disk. NOTE If the error occurs again, please contact your dealer for assistance. 8-21 8 Troubleshooting 8.2.7 Temperature Controller Errors Error code E0565 E1500 8 Error message ISE temperature control ERR Problem with ISE. Contact your dealer. When the measurement results are appended with a “+” or “-” mark or an error listed to the left occurs, the Temperature control temperature inside the incubator is outside the ERR specified range. Check the ambient temperature. Clean the air filter. Reboot the analyzer. E1501, E1505 Temperature control ERR Check the ambient temperature. Reboot the analyzer. E1510 ISE temperature control ERR Check the ambient temperature. Clean the air filter. Reboot the analyzer. E1511, E1515 Temperature control ERR Check the ambient temperature. Reboot the analyzer. E1520, E1521 Room temperature WARNING Check the ambient temperature. Reboot the analyzer. 8-22 Error condition Troubleshooting Please contact your dealer for assistance. Perform the following troubleshooting. (a) Check the room temperature is within 15 to 32 °C. If the temperature is out of the range, control the room temperature. (b) Switch off the power. (c) Clean the air filter. è Refer to Section 6.2 (d) Make sure that thumbscrews of the incubator are tightened properly and the incubator cable is connected to the connector properly. èRefer to Section 6.3.3 (e) Make sure that the sample application unit cover is attached to the analyzer properly. Tighten the thumbscrews securely. (f) Leave the analyzer for about 10 minutes to allow the interior to cool, and then switch on power. NOTE If the error occurs again, please contact your dealer for assistance. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting Error code E1502 Error message Error condition Temperature control ERR Incubator temperature sensor disconnected. Check the incubator cable. Contact your dealer. When the measurement results are appended with a “$” mark or an error listed to the left occurs, the incubator cable either is not connected or has been severed. E1503 Temperature control ERR Incubator temperature sensor damaged. Contact your dealer. E1504 Temperature control ERR Incubator heater disconnected. Check the incubator cable. Contact your dealer. E1512 ISE temperature con- Disconnection error of trol ERR the ISE incubator ISE temperature control element damaged. Contact your dealer. E1514 ISE temperature control ERR ISE heater disconnected. Contact your dealer. 897N203029A FDC NX600 Instruction Manual 2022.01 Troubleshooting Switch off the power and connect the incubator cable again. NOTE After inserting the connector of the incubator cable, you must turn it to the right to lock it. è Refer to Section 6.3.3 NOTE If the error occurs again, please contact your dealer for assistance. Please contact your dealer for assistance. 8 8-23 8 Troubleshooting 8.2.8 Circuit Board Signal Errors Error code 8 Error message Error condition E0035 Light too intense A circuit board signal error has occurred. Light too intense. Contact your dealer in case the ERR repeats. E0300, E0301, E0302 Gain setting ERR Contact your dealer in case the ERR repeats. E0509 AD conversion calibration ERR Contact your dealer. E0510 Faulty A/D controller Contact your dealer. E0515 DA conversion initialization ERR Contact your dealer. E0516 AD conversion reset ERR Contact your dealer. E0517 AD conversion initialization ERR Contact your dealer. E0532, E0533, E0537, E0538 ISE self test AD conversion ERR For ISE measurement, turn the power SW OFF and ON. Contact your dealer in case the ERR repeats. Other parameters besides ISE can be measured. E0534, E0535, E0536, E0539 ISE AD conversion ERR For ISE measurement, turn the power SW OFF and ON. Contact your dealer in case the ERR repeats. Other parameters besides ISE can be measured. 8-24 Troubleshooting If the error still occurs after switching the analyzer off and on, please contact your dealer for assistance. Please contact your dealer for assistance. When an error occurs after turning the power on and off, please contact your dealer for assistance. Tests other than ISE do not require a restart. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting Error code Error message Error condition Troubleshooting E0540 ISE AD conversion A circuit board signal initialization ERR error has occurred. For ISE measurement, turn the power SW OFF and ON. Contact your dealer in case the ERR repeats. Other parameters besides ISE can be measured. When an error occurs after turning the power on and off, please contact your dealer for assistance. Tests other than ISE do not require a restart. E1530, E1531, E1532 Temperature AD conversion ERR Contact your dealer. Please contact your dealer for assistance. E1750E1755, E1757 Slide reading unit failure Contact your dealer. If the error still occurs after switching the analyzer off and on, please contact your dealer for assistance. E3900 Internal communication ERR Contact your dealer. E4401, E4403 Save ERR Contact your dealer. E4410 Measurement result reference error The number of measurement results, with the new result, currently exceeds capacity. The selected measurement result was deleted. This measurement result cannot be referred. Control chart reference error The selected chart is not up-to-date with the new measurement. Please go back to the parameter selection screen and display the chart again. Please go back to the parameter selection screen and display the chart again. E4411 897N203029A FDC NX600 Instruction Manual 2022.01 8 8-25 8 Troubleshooting Error code 8 Error message Error condition Troubleshooting E4454 Memory loading ERR A circuit board signal error has occurred. Contact your dealer. W4460 W4461 W4464 W4469 W4470 W4473 W4474 Memory recovery WARNING (Example: setting info.) Recovery performed from backup memory. Contact your dealer. W4462 W4463 Memory recovery WARNING Recovery performed from backup memory. Check whether the setting for reference intervals is correct. Check whether the setting for reference intervals is correct. W4465 W4466 W4468 Memory recovery WARNING Recovery performed from backup memory. Check whether the QC card is read by confirming the lot number. If not, re-read the QC card. Check the lot information whether the QC card has been read. If the card has not been read, read the QC card again. W4467 Memory recovery WARNING Recovery performed from backup memory. Check whether the QC card is read by confirming the lot number. If not, contact your dealer. Check the lot information whether the QC card has been read. If the card has not been read, contact your dealer for assistance. W4471 W4472 Memory recovery WARNING Recovery performed from backup memory. Check the control fluid setting. Check the setting for control fluids. 8-26 Please contact your dealer for assistance. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting Error code Error message Error condition Troubleshooting E4475, E4478E4481 Memory initial-value A circuit board signal recovery ERR error has occurred. Recovery performed from backup memory. Contact your dealer. Please contact your dealer for assistance. W4477 Memory recovery WARNING Recovery performed from backup memory. Check the Operator ID setting. Check the operator ID setting. W4483 Memory recovery WARNING (Password) The administrator password has been returned to the default. Contact your dealer in case the ERR repeats. If the error occurs again, please contact your dealer for assistance. E4490 File deletion ERR Contact your dealer in case the ERR repeats. Please contact your dealer for assistance. E4600E4613 Calculation ERR Contact your dealer. W4720 Clock setting WARN- The date and time ING have been reset. Set date & time. Contact your dealer in case the ERR repeats. Set the date and time using [Date & Time settings] in Normal mode. è Refer to Section 9.2.1 If the error occurs again, please contact your dealer for assistance. IMPORTANT: If the date and time are not adjusted correctly, the analyzer may fail to determine the expiration of the slides and the calibration, and the measurement results may not be accurate. 897N203029A FDC NX600 Instruction Manual 2022.01 8-27 8 8 Troubleshooting Error code 8 Error message Error condition Troubleshooting W4721 Clock control failure A circuit board signal WARNING error has occurred. Set date & time. Contact your dealer in case the ERR repeats. Set the date and time using [Date & Time settings] in Normal mode. è Refer to Section 9.2.1 W4733 Power switch control WARNING Contact your dealer. Please contact your dealer for assistance. W5000 USB memory read ERR Contact your dealer. W5002W5019 USB memory read USB memory read erERR ror has occurred. Lack of necessary files in the USB memory. Check the USB memory, and then reselect a menu. Contact your dealer in case the ERR repeats. (Failed to copy files from the USB memory.) Check the USB memory and retry according to the error message. If the error occurs again, please contact your dealer for assistance. W5020 USB memory write A circuit board signal ERR error has occurred. Contact your dealer. Please contact your dealer for assistance. W5050 No USB memory USB memory can not be recognized. Insert USB memory. If the error occurs again, please contact your dealer for assistance. E5500 A b n o r m a l s a m p l e Signal system error spreading on slide inside circuit board. Contact your dealer. When an error occurs after turning the power on and off, please contact your dealer for assistance. E5901 Software logic ERR Contact your dealer. Please contact your dealer for assistance. 8-28 The USB memory is not inserted. A software error has occurred. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.9 QC Card, PF Card, and DI Card Errors (1) When data is not displayed after a QC card is read Error code Error message Error condition Problem reading QC card information ― Troubleshooting Read the QC card again after making sure its correct orientation. If it does not improve, insert the air blow nozzle into the QR code reader slot and blow out the dust inside. ― Spray by pushing in the end of the nozzle about 50 mm. (2) When QC information has not been read Error code Error message Error condition 8 Troubleshooting W0170 No QC card info No QC information. Read QC card. The QC card of the slides loaded in the slide cartridge has not been read. Read the QC card of the printed slide code. W4010 No QC type info. No type information corresponding to the slide. Read QC card corresponding to the slide. There is no type information for the corresponding item. Read the corresponding QC card. (3) When a PF card has not been read Error code W0137 Error message No PF card info. No PF card info. Read PF card. Error condition Troubleshooting The PF sample rack Read the PF card included with the PF. has been identified, but no PF card has not been read. 897N203029A FDC NX600 Instruction Manual 2022.01 8-29 8 Troubleshooting (4) When a DI card has not been read Error code W4012 Error message Error condition No DI info. There is no DI card No DI info found. Can- information. not read this card. Contact your dealer. Troubleshooting Please contact your dealer for assistance. (5) Other precautions Error code Error message Error condition Troubleshooting W1700 QC card reader failure The card reader has QC card reader dam- malfunctioned. aged. Measurement is still possible. Contact your dealer. Please contact your dealer for assistance. W1703 QC card read error QC card read ERR Check the QC card direction and re-read it. (Insert the QR code so that it faces the far left side.) If the error occurs repeatedly, contact your dealer. Make sure the card is inserted at correct orientation. W4011 QC card ERR Cannot save data. Panel card read error Please contact your dealer for assistance. W4013 QC card ERR Check whether this is the correct QC card. A QC card data error has occurred. Check whether the QR code of the QC card is dirty. 8 NOTE When the error is not resolved, please contact your dealer for assistance. W4014 QC card ERR The QC card cannot Cannot read QC card be read at this time. for “Mode Function Underway” or “Spotting Underway”. Read again after mode function ends, or after spotting ends. After mode function or sample application ends, read the card again. W4015 Expired QC card QC card has expired and cannot be used. Use QC cards that are before their expiration dates. 8-30 QC card has been expired. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.10 Slide Loading Errors Error code Error message Error condition Troubleshooting W0040 Incorrect sample WARNING The sample type set in analyzer does not correspond to the read sample type. Set the sample type by SAMPLE key and the slide’s sample type corresponding to the sample. Rerun the tests from the beginning. The sample type selected by the [Sample type] button and the sample type of the slide do not match. (Example) • Whole blood (W) slides were loaded when either [P/S] or [U] had been selected by the [Sample type] button. • Plasma/serum (P) slides were loaded when [W] had been selected by the [Sample type] button. Match the configuration by the [Sample type] button and the sample type to measure the slide to be used, and then rerun the measurement from the beginning. W0041 Incorrect sample type The sample type setWARNING ting of PF measurement does not match. Although a PF sample rack is loaded, the sample type setting is not P/S. Set the sample type to P/S. W0042 Incorrect sample rack WARNING Correct sample rack not set in place. Check the sample rack. Load a PF sample rack in which a sample extracted by PF has been loaded. A re-run test for a measurement by using PF is attempted, but a rack other than a PF sample rack was loaded. 8 A PF sample rack has Load a sample rack other than a PF been loaded for a con- sample rack. trol measurement. A sample rack other Load a sample rack for calibrator CP. than for CP has been loaded for a calibration measurement. W0043 N o s a m p l e r a c k No sample rack is WARNING loaded. Sample rack is not set. Set a sample rack. W0090 ISE slide direction The ISE slide is loaded Load the ISE slide in the slide cartridge in reverse. in the proper direction, and then measERR ure again. Incorrect loading direction of ISE slide. Check the ISE slide. 897N203029A FDC NX600 Instruction Manual 2022.01 Set a sample rack. 8-31 8 Troubleshooting Error code Error message Error condition Troubleshooting E0160 Incorrect diluent type Diluent type does not error match the slide to be used. Diluent type is incorrect. Use specified diluent for the test name. W0175 No slide Slides have not been set. Rerun the tests from the beginning. No slides are loaded in Load slides to be tested, and rerun the the slide cartridge. tests from the beginning. W5400 Expired slide Slide is expired and cannot be used. The validity term of the Always use valid slides. slide has expired. Use specified diluent for the test name, and rerun the measurement from the beginning. IMPORTANT: Slide validity may not be detected in ISE tests. è Refer to Section 8.1.3 NOTE Depending on the device settings, this warning may not be generated. 8 8-32 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.11 ISE Measurement Errors IMPORTANT: ISE measurement values to which error codes have been appended (E1 - E8) may not be correct. Perform the following troubleshooting and rerun the measurement. (1) When a self-test error occurred Error code Error message Error condition E0530 ISE self test gain ERR For ISE measurement, turn the power SW OFF and ON. Contact your dealer in case the ERR repeats. Other parameters besides ISE can be measured. An error was detected during a self-test performed before the start of an ISE measurement. E0531 ISE self test offset ERR For ISE measurement, turn the power SW OFF and ON. Contact your dealer in case the ERR repeats. Other parameters besides ISE can be measured. 897N203029A FDC NX600 Instruction Manual 2022.01 Troubleshooting The analyzer must be restarted for an ISE measurement. If the error occurs again, please contact your dealer for assistance. Tests other than ISE do not require a restart. 8 8-33 8 Troubleshooting (2) Errors are appended to measurement results Check menu Description Drift check Checking abnormal time course (voltage) during testing Impedance check Checking the slide impedance after completing a test Over range check Checking test result is within the measurement range Code: During an ISE measurement, the checks listed in the table to the left are performed. If there is a problem with a measurement value, an error code from E1 through E8 is appended to the measurement result. Description E1 Drift error E2 Impedance error E3 Outside of the measurement range (This is not an analyzer malfunction.) E4 Drift error and impedance error E5 Impedance error and outside of the measurement range E6 Drift error and outside of the measurement range E7 Drift error, impedance error, and outside of the measurement range E8 Impossible measurement < When the error code is not E3 > The sample or reference fluid may have not been sufficiently absorbed into the slide. Perform the following troubleshooting steps. 8 1. Inspect the sampler O-ring. è Refer to Section 6.6.2 2. Clean the spotting part and ISE probe unit. è Refer to Section 6.3.2 3. Make sure that the analyzer can measure correctly. è Refer to Chapter 7 NOTE If the error occurs again, please contact your dealer for assistance. < When the error code is E3 > There is no problem with the analyzer. The concentration of the sample is outside of the measurement range. NOTE Refer to the “Instructions for Use” of the slide for the measurement range. 8-34 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting (3) Other Errors Error code Error message Error condition Troubleshooting W0091 Unacceptable ISE test ISE tests cannot run. Turn the power SW OFF and ON. Rerun the tests from the beginning. An ISE measurement error has occurred, and ISE measurement cannot be started. Check ISE measurement errors that occurred before this error, and then carry out the corresponding troubleshooting. After troubleshooting, turn the power of the device on and off, and then start measurement again from the beginning. W0092 Unacceptable ISE manual pipetting ISE test cannot be done in this test mode. Set a measurable slide and redo the test from the beginning. There was an attempt to measure an ISE slide using manual sample application measurement. Measure an ISE slide using automatic sample application. 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-35 8 Troubleshooting 8.2.12 PF Errors Error code 8 Error message Error condition Troubleshooting E0132 PF pressure ERR Replace the PF pump tube. A pressure error ocReplace the PF Tube. curred during the filter- è Refer to Section 6.9.4 ing process. E0133, E0150 PF sensor ERR Contact your dealer. A PF sensor error has occurred. W0134 PF sampling ERR Error occurred during PF sampling. Check for a PF leak. An error occurred dur- Clean the PF Suction Pad. ing PF filtering. è Refer to Section 6.9.1 W0138 Unacceptable PF sampling WARNING For PF measurement, turn the power SW OFF and then ON. Contact your dealer in case the ERR repeats. Measurements other than PF measurements can be measured without turning the power SW OFF and then ON. A PF measurement error has occurred, and PF measurement cannot be started. Check PF measurement errors that occurred before this error, and then carry out the corresponding troubleshooting. After troubleshooting, turn the analyzer on and off, and then start measurement again from the beginning. E1300, E1301, E1303, E1304 PF motor ERR PF motor error moveProblem with the plas- ment error ma filter. Contact your dealer. Turn the power on and off. If the error occurs again, please contact your dealer for assistance. W1308 PF sensor ERR A PF sensor error has Problem with the plas- occurred. ma filter. Contact your dealer. Please contact your dealer for assistance. E1302 PF motor operation PF motor error moveERR ment error Problem with the plasma filter. Contact your dealer. Turn the power on and off. If the error occurs again, please contact your dealer for assistance. 8-36 Please contact your dealer for assistance. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.13 Errors Related to Data Communication, Sample Barcode Reader (1) When a communication error occurs Error code W4500, W4501, W4502 Error message Error condition Serial communica- Serial communication tion ERR error Check the specifications and connection of the communication cable. Contact your dealer in case the ERR repeats. Troubleshooting Check the following: communication cable, host computer, and communication destination settings. è Refer to Section 9.3.2 NOTE If the error occurs again, please contact your dealer for assistance. W4510 LAN communication LAN communication ERR error Check the specification and the connection of the communication cable. Contact your dealer in case the ERR repeats. W4511 Type 4 LAN commu- Type 4 LAN communi- Check the host computer and type 4 nication ERR cation ERR settings. Check whether the setè Refer to Section 9.3.2 ting for type 4 or host NOTE If the error occurs again, please computer is correct. contact your dealer for assisContact your dealer in tance. case the ERR repeats. W4520 Incorrect work list in- The received work list formation information does not Check whether work list match specifications. information for the host computer is correct. Contact your dealer in case the ERR repeats. E6200, E6201, E6210 Check the work list information. NOTE If the error occurs again, please contact your dealer for assistance. Communication ERR A communication error Please contact your dealer for assishas occurred. tance. Contact your dealer. 897N203029A FDC NX600 Instruction Manual 2022.01 8-37 8 8 Troubleshooting (2) Sample barcode reader error Error code Error message Error condition Troubleshooting W5330 Barcode reader not The sample barcode Connect the sample barcode reader to connected reader is not connected. the USB connector, and then turn the analyzer off and on. Check the barcode reader connection. W5331 Exceeded number of Barcode label error characters from barcode reader The number of input characters from the bar code reader exceeded the capacity. Check the barcode label. W5332 Barcode WARNING Barcode data error Check input value and re-enter. Check the input value and re-enter. 8 8-38 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting 8.2.14 Calibration Errors (1) An immuno-chemical test has been performed without calibration. Error code W0180 Error message Error condition Troubleshooting Calibration not per- An immuno-chemical Perform the calibration. formed test has been perè Refer to Section 4.1 formed without calibraPerform calibration. tion. (2) A slide not suited for calibration has been loaded. Error code Error message Error condition Troubleshooting W0190 Wrong slide for calibration A slide that does not require calibration is loaded. Check the slide and perform the calibration from the beginning. A slide not suited for Check the slide. Rerun the calibration calibration has been process again. loaded during the calibration process. W0191 Wrong slide lot for calibration A different lot of slide other than designated in the calibration mode is loaded. Load a slide corresponding to the designated lot and perform the calibration from the beginning. A slide different from Load a slide of the specified lot, and the lot specified during then calibrate again. calibration has been loaded. 8 (3) Insufficient calibration data Error code E4200 Error message Calibration ERR Invalid number of slides The number of measured slides for the following calibration data is insufficient or the sample is unspotted. Error condition Troubleshooting This error may occur Rerun 2 measurements for the printed due to a lack of calibra- level. tion data or no application on the slide. 897N203029A FDC NX600 Instruction Manual 2022.01 8-39 8 Troubleshooting (4) Calibration result problem Error code E4201 Error message Error condition Troubleshooting Calibration variation The variation of meas- Rerun 2 measurements for the printed error urement values during level. Variations in measured calibration is large. value for the following calibration are large. (5) Standard curve configuration error Error code Error message Error condition E4202 Calibration curve set- The wrong calibrator ting ERR (span error) may have been used. Misplacement of calibrator. Check the level and expiration date of the calibrator and perform the calibration from the beginning. E4203 Calibration curve setting ERR (range error) Misplacement of calibrator. Check the level and expiration date of the calibrator and perform the calibration from the beginning. E4205 Calibration calcula- Calibration calculation tion ERR problem Failed in calibration calculation. Do the calibration again. 8 8-40 Troubleshooting Check the expiration date and the setting position of the calibrator, and then calibrate again from the beginning. Calibrate again from the beginning. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting (6) Measurement errors during calibration Error code E4204 Error message Calibration measurement ERR Error in measured result. After coping with the printed error, perform the calibration from the beginning. Error condition Error marks such as “ * ”, “ ? ”, and “$” are printed on the measurement results. Troubleshooting Troubleshoot the printed errors and then calibrate again from the beginning. 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-41 8 Troubleshooting 8.2.15 Other Errors Error code 8 Error message Error condition Troubleshooting W0020 No disposal box Measurement was started with the disPut Disposal box to the end and then press posal box pulled out. START key. Insert the disposal box completely, and then start measurement. E0021 Disposal box open ERR Disposal box opened during measurement. Close Disposal box and run the tests from the beginning. Close the disposal box, and then start measurement again from the beginning. W0022 Disposal box full The disposal box is full Disposal box became of slides and tips. full of slides and tips. Discard slides and tips to empty the box when measurement is not in progress. While a measurement is not being performed, discard the slides and tips in the disposal box. E0080 Sample tip sensor ERR Contact your dealer. Malfunction of the sample tip sensor. Please contact your dealer for assistance. W0500 Rear fan 1 stop Contact your dealer. A fan error has occurred. W0501 Rear fan 2 stop Contact your dealer. W0502 Intake fan stop Contact your dealer. W0503 Exhaust fan stop Contact your dealer. I5920 Turn the power ON A long time has passed Switch the analyzer off and on. 30 days or more have since start-up. passed from when the analyzer was started. For stable system performance, turn the power ON. 8-42 The disposal box was open during measurement. NOTE When there are 150 slides or 75 tips in the disposal box, this error occurs. If the disposal box is pulled out, it is determined that the slides and tips were discarded, and the disposal count is reset. Always empty the disposal box after pulling it out. 897N203029A FDC NX600 Instruction Manual 2022.01 8 Troubleshooting Error code Error message Error condition Troubleshooting W4440 Setting value reset The settings have WARNING been reset. The reference interval range setting, control fluid setting, and correlation coefficients have been reset. Check the settings. Check settings for the reference interval range and the control fluid, and check correlation coefficients. W66xx The entered value is Input value ERR Check input value and incorrect. re-enter. Check the entered value and enter it again. 8 897N203029A FDC NX600 Instruction Manual 2022.01 8-43 8 Troubleshooting 8.3 Slide Code Table Slide code Test Test code Sample code Slide code Test Test code Sample code GLU -P 10 50 GGT -P 30 50 BUN -P 11 50 GOT/AST -P 31 50 UA -P 13 50 GPT/ALT -P 32 50 TCHO -P 14 50 CPK -P 33 50 NH3 -P 15 50 LDH -P 34 50 TG -P 16 50 ALP -P 35 50 CRE -P 17 50 LAP -P 37 50 TP -P 18 50 CKMB -P 38 50 ALB -P 20 50 CHE -P 39 50 TBIL -P 21 50 AMYL -P 43 50 Ca -P 23 50 LIP -P 44 50 IP -P 24 50 TCO2 -P 45 50 DBIL -P 25 50 CRP -S 51 50 HDLC -P 26 50 Na 91 00 Mg -P 28 50 K 92 00 Cl 93 00 GLU -W 10 00 NH3 -W 15 00 NOTE Slide codes are subject to change without notice. Check the slide code printed on the slide box. 8 8-44 897N203029A FDC NX600 Instruction Manual 2022.01 9 If the Function Menu (Function menu) button is tapped, the Function menu screen appears. Function menu list Function menu items Settings Advanced settings Calibration — — Reference section 4.1.1 Control — — 9.1 START Manual pipetting — — 4.2 Date & Time settings — 9.2.1 Brightness • Volume Display brightness 9.2.2 Error sound Other general sound Lamp management Lamp ON time 9 9.2.3 Lamp off timer setting Lot information Normal mode Sampler leak check — Sample nozzle 9.2.4 6.6.2 Reference fluid nozzle Replace O-ring PF maintenance — 6.6.3 Cleaning 6.9.1 Leak check 6.9.2 Suction pad 6.9.3 Reference plate level check — 6.3.1 Cleaning mode — 6.3.2 897N203029A FDC NX600 Instruction Manual 2022.01 9-1 9 Function Menu Function menu items Settings Advanced settings Language setting — Host connection setting None Reference section 9.3.1 9.3.2 Serial connection LAN connection Calculated Parameter — 9.3.3 Unit conversion — 9.3.4 Switch measurement range display — Reference interval range setting — 9.3.6 Sample No. /ID settings — 9.3.7 Work list selection setting Sample ID / Patient name 9.3.5 9.3.8 Patient ID / Patient name Mode function for administrator Correlation coefficients — 9.3.9 Lot compensation coefficients — 9.3.10 Standard curve coefficients — 9.3.11 Lot Qualification — 9.3.12 PF calibration coefficients — 9.3.13 Dilution setting — 9.3.14 Tube setting for diluent & reference fluid Diluent 9.3.15 Reference fluid Print setting for measured result Number of printing sheets 9.3.16 Print order Print with the ranges Control fluid setting — 9.3.17 Control measurement setting — 9.3.18 Operator ID setting — 9.3.19 Test item name setting 9 Display and printout 9.3.20 Data transmission UREA setting Error log 9-2 Default sample type setting — 9.3.21 Password setting — 9.3.22 — — 8.1.1 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.1 Control This mode is used for measurement of control serum (FUJI DRI-CHEM CONTROL QP) and for performing control surveys. When control mode is set, measurements are performed under the conditions listed below. • Correlation coefficients: follow [Control measurement setting] (Initial values disabled (a=1, b=0)) of the function menu. • Measurement units: follow [Unit settings] of the function menu. • Dilution conditions: dilution is not normally performed. (This can be configured at the Sample information registration screen.) NOTE When the analyzer is switched on again, control mode is canceled. 1. Prepare the slide to be measured, and then load it on the analyzer. 2. Place the control fluid to be measured in the tube, and then load it in the sample position. è Refer to Section 3.3 (3) 3. Tap the button. 4. Tap the [Control] button. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-3 9 Function Menu 5. Enter or choose the sample information. 1 2 3 4 5 6 1 Enter the sample number using the software keyboard (Section 2.3). 2 Enter the sample ID using the software keyboard (Section 2.3). 3 Select the control liquid from the list (when necessary). 4 Select the control liquid lot from the list (when necessary). IMPORTANT: 9 When you configure the control liquid lot, make sure it matches the lot of the control liquid to be used. If it does not match, the measurement may not be valid. 5 For the sample type, generally (in the case of QP-L, QP-H, and QE) select [Plasma/Serum]. 6 For dilution, generally select [None]. 9-4 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sampler cover can be opened. 7. Tap the [START] button to start control measurement. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-5 9 Function Menu 8. After measurement, tap the button to check the measurement results. IMPORTANT: After using control mode, always tap the [HOME] button to clear control mode. <Display of measurement results> Green: when within the acceptable range of control Red: when outside of the acceptable range of control Acceptable lower limit Acceptable upper limit NOTE The acceptable range of control displays the value entered for [Control fluid setting] (see Section 9.3.17). 9 9-6 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.2 Normal Mode General operators use this function menu to, for example, set the time and configure the brightness of the LCD. Settings Advanced settings Date & Time settings Brightness • Volume — Display brightness Reference section 9.2.1 9.2.2 Error sound Other general sound Lamp management Lamp ON time 9.2.3 Lamp off timer setting Lot information Sampler leak check — Sample nozzle 9.2.4 6.6.2 Reference fluid nozzle Replace O-ring PF maintenance — 6.9.1 Leak check 6.9.2 Suction pad 6.9.3 Reference plate level check — 6.3.1 Cleaning mode — 6.3.2 897N203029A FDC NX600 Instruction Manual 2022.01 9 6.6.3 Cleaning 9-7 9 Function Menu 9.2.1 Date & Time Settings This mode is used to set the date and time settings stored in the memory. • Tap “ ” or “ ” to change the date and time values. • Configure the order of the year, month, and date by using [Date format setting]. NOTE The analyzer must be restarted to enable a change of [Date format setting]. The confirmation dialog for the poweroff will appear when is tapped to exit Normal mode. After the shutdown, reboot the analyzer. 9.2.2 Brightness • Volume You can configure the brightness of the LCD and buzzer sounds (error warning, measurement completed). 9 • Configure the brightness of the LCD using and to set the [Display brightness]. Use minimum and maximum brightness. • Configure the buzzer sound when an error ocand to curs by using [Error sound]. Use set the minimum and maximum. • Configure the buzzer sound when measurement ends by using [Other general sound]. Use and to set the minimum and maximum. If is configured, the buzzer is muted. 9-8 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.2.3 Lamp Management Display and reset the cumulative illumination time of the lamp currently installed in the analyzer. Furthermore, choose whether to turn off the lamp automatically when the analyzer has not been used for some time. • You can check the cumulative illumination time by using [Lamp ON time]. If the [Reset] button is tapped, the cumulative illumination time is reset. NOTE Reset the cumulative illumination time of the lamp after replacing the lamp. • If the [Lamp off timer setting] checkbox is selected, the lamp is switched off automatically after some time. You can configure the time until the lamp is switched off. NOTE The default setting is 20 minutes. 9.2.4 Lot Information The screen displays the lot information of currently registered QC cards. You can also print all currently registered lot information. • When there are two or more pages that can be displayed, tap the and buttons to scroll the pages. • If the button is tapped, all currently registered lot information is printed. NOTE Up to five lots are registered for one test. 897N203029A FDC NX600 Instruction Manual 2022.01 9-9 9 9 Function Menu 9.3 Mode Function for Administrator NOTE To display Mode function for administrator, open the Function menu page, tap the [Mode function for administrator] button, and then enter the password. NOTE In order to prevent other people from operating the analyzer, the input dialog for the password will appear when no operation is performed for a fixed period of time during the operation of [Mode function for administrator]. Settings Language setting Host connection setting 9 Calculated Parameter Unit conversion Switch measurement range display Reference interval range setting Sample No. /ID settings Work list selection setting Correlation coefficients Lot compensation coefficients Standard curve coefficients Lot Qualification PF calibration coefficients Dilution setting Tube setting for diluent & reference fluid Print setting for measured result Control fluid setting Control measurement setting Operator ID setting Test item name setting Default sample type setting Password setting 9-10 Advanced settings — None Serial connection LAN connection — — — — — Sample ID / Patient name Patient ID / Patient name — — — — — — Diluent Reference fluid Number of printing sheets Print order Print with the ranges — — — Display and printout Data transmission UREA setting — — Reference section 9.3.1 9.3.2 9.3.3 9.3.4 9.3.5 9.3.6 9.3.7 9.3.8 9.3.9 9.3.10 9.3.11 9.3.12 9.3.13 9.3.14 9.3.15 9.3.16 9.3.17 9.3.18 9.3.19 9.3.20 9.3.21 9.3.22 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.1 Language Setting Configure the language to be used. Tap the button of language to be configured. NOTE Rebooting the device is necessary for the setting to take effect. The confirmation dialog for the power-off will appear when is tapped to exit Mode function for administrator. After the shutdown, reboot the analyzer. 9.3.2 Host Connection Setting Configure the connection terminal and communication parameters to connect to the host computer (PC). Furthermore, switch the built-in printer on or off. • When you will not connect to a host computer, set [None]. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-11 9 Function Menu • If [Serial connection] is set, configure the communication type and whether to output from the printer. You can choose from 3 communication types. - Type 1: Bidirectional connection - Type 2: Unidirectional connection (bidirectional format) - Type 3: Unidirectional connection (standard format) NOTE Rebooting the device is necessary for the setting to take effect. The confirmation dialog for the power-off will appear is tapped to exit Mode function when for administrator. After the shutdown, reboot the analyzer. • If [LAN connection] is set, configure the communication type and whether to output from the printer. You can choose from 4 communication types. - Type 1: Bidirectional connection - Type 2: Unidirectional connection (bidirectional format) - Type 3: Unidirectional connection (standard format) - Type 4: Unidirectional connection (HL7 formatted) was tapped, the LAN configuration • When screen appears. If an entry field is tapped, the software keyboard (refer to Section 2.3) appears so you can edit the value. 9 NOTE The following section shows the default LAN settings. FDC NX600 settings IP address:192.168.100.1 Subnet mask:255.255.255.0 Port number:65535 Gateway: 0.0.0.0 PC settings IP address:192.168.100.2 Port number:65535 NOTE Rebooting the device is necessary for the setting to take effect. The confirmation dialog for the power-off will appear is tapped to exit Mode function when for administrator. After the shutdown, reboot the analyzer. 9-12 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu If the button is also tapped when Type 4 was selected, the settings screen of Type 4 of page 3 becomes active. • Tap [Examination code] to select the standard for examination item classification code to be used in communication. • Tap [MLLP setting] to set MLLP (Minimal Lower Layer Protocol) in communication. MLLP is applied to full message in comFull: munication. Custom: MLLP is applied to only at the end of message in communication. • Tap [Sample type] to select the sample type used in communication. NOTE Rebooting the device is necessary for the setting to take effect. The confirmation dialog for the power-off will appear when is tapped to exit Mode function for administrator. After the shutdown, reboot the analyzer. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-13 9 Function Menu 9.3.3 Calculated Parameter You can configure the settings to calculate the following items from simultaneously measured values and then display the results. Items registered for calculation have their measurement results displayed and printed. Test name LDL Cholesterol (LDL-C) non-HDL Cholesterol (non HDL-C) Globulin Calculation test name LDL non-HDL GLOB Albumin/Globulin ratio ALB/GLOB Unit Calculation formula LDL-C = TCHO value - (HDL-C value + TG value/5) [mg/dL] mg/dL LDL-C = TCHO value - (HDL-C value + TG value/2.2) [mmol/L] or mmol/L NOTE LDL-C is NOT calculated when the measured value of TG exceeds 400 mg/dL (4.52 mmol/L). mg/dL non-HDL = TCHO value - HDL-C value or mmol/L g/dL or GLOB = TP value - ALB value g/L — ALB/GLOB = ALB value / (TP value - ALB value) BUN/CRE = BUN value/CRE value [mg/mg] BUN/CRE = BUN value ×1,000/CRE value[(SI)] BUN/Creatinine ratio GOT/GPT ratio (AST/ALT ratio) Sodium/Potassium ratio 9 Anion Gap BUN/CRE GOT/GPT (AST/ALT) mg/mg NOTE BUN measured value in SI unit (mmol/L) is or shown as molar concentration of Urea. BUN/ (SI) CRE ratio in SI unit (molar ratio) is shown Urea concentration/creatinine concentration ratio. GOT/GPT = GOT value / GPT value (AST/ALT = AST value / ALT value) — NOTE The calculation test name can be changed by [Test item name setting] in the Mode function for administrator. è Refer to Section 9.3.20. Na/K — Na/K = Na value/K value mEq/L Anion Gap = Na value - (Cl value + TCO2 value) Anion Gap or mmol/L IMPORTANT: Calculation items display a value calculated from simultaneously measured values. Therefore, calculation results are impacted by errors included in measurement values and the effects of interfering materials. Please understand that when you use the results for a diagnosis, you must make an overall determination in combination with the measurement values is used in the calculation. NOTE In the case of additional testing and retesting, items subject to calculation must be measured at the same time. NOTE When there are two or more slides of the same item subject to calculation in the slides measured at the same time, the initial measurement value is used. NOTE When the measurement results of items subject to calculation exceed the measurement range (determination range), the calculation results of this calculation item are displayed and printed as "****". NOTE When an item subject to calculation has expired, an (expiry) warning symbol (#) is displayed and printed for the measurement results of this item and the calculation results. 9-14 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu If the [Calculated Parameter] button is tapped in Mode function for administrator, precautions for use of calculation items are displayed. To use calculation items after confirming the precautions, tap the [OK] button. Configure the calculation items to be used. NOTE If the QC card of an item to be used for calculation has not been read, the calculation item is not displayed. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-15 9 Function Menu 9.3.4 Unit Conversion Configure the main unit and secondary unit for each measurement item. The following table shows the units that can be used for each configuration item and the conversion coefficient Test Test code Unit (A) Unit (B) Conversion coefficient GLU 10 mg/dL mmol/L 0.05551 BUN 11 mg/dL UA 13 TCHO Test Test code Unit (A) Unit (B) Conversion coefficient GGT 30 U/L μkat/L 0.0167 mmol/L 0.357 GOT/ AST 31 U/L μkat/L 0.0167 mg/dL μmol/L 59.48 GPT/ALT 32 U/L μkat/L 0.0167 14 mg/dL mmol/L 0.02586 CPK 33 U/L μkat/L 0.0167 NH3 15 μg/dL μmol/L 0.7139 LDH 34 U/L μkat/L 0.0167 TG 16 mg/dL mmol/L 0.01129 ALP 35 U/L μkat/L 0.0167 CRE 17 mg/dL μmol/L 88.4 LAP 37 U/L μkat/L 0.0167 TP 18 g/dL g/L 10 CKMB 38 U/L μkat/L 0.0167 ALB 20 g/dL g/L 10 CHE 39 U/L μkat/L 0.0167 TBIL 21 mg/dL μmol/L 17.1 AMYL 43 U/L μkat/L 0.0167 Ca 23 mg/dL mmol/L 0.25 LIP 44 U/L μkat/L 0.0167 IP 24 mg/dL mmol/L 0.3228 TCO2 45 mmol/L DBIL 25 mg/dL μmol/L 17.1 CRP 51 mg/dL HDL-C 26 mg/dL mmol/L 0.02586 Na 91 mEq/L mmol/L 1 Mg 28 mg/dL mmol/L 0.4114 K 92 mEq/L mmol/L 1 Cl 93 mEq/L mmol/L 1 mmol/L 1 mg/L 10 NOTE Unit (B)= Unit (A)× Conversion coefficient NOTE BUN measured value in Unit (B) (mmol/l) is shown as molar concentration of Urea. • Tap the button of the unit to be used for each measurement item. 9 Next-page buttons 9-16 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.5 Switch Measurement Range Display Switch the display method for data outside the measurement range (passing the upper/lower limit of the determination range) for all tests. • If [@] is configured, an [@] symbol is added to indicate the displayed measurement value is outside the measurement range. • If [< / >] is configured, a value outside the measurement range is displayed by a [<] or [>] symbol. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-17 9 Function Menu 9.3.6 Reference Interval Range Setting You can add, edit, and change the order of the reference interval names that can be selected when registering sample information. (Refer to Section 3.3 (4)) Furthermore, you can set a reference interval for each sample type ([P/S, W] or [U]) of a reference interval name. Adding and editing of reference interval names Tap the add. button of the item you want to edit or Next-page buttons If an entry field is tapped, the software keyboard (refer to Section 2.3) appears so you can edit the test name and other details. To change the sample type when configuring the reference interval, tap the [P/S, W] button or [U] button. 9 9-18 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu Changing the display order Tap the button of the test for which you want to change the display order. Next-page buttons Tap the field of the test for which you want to configure the transfer destination, and then tap the [OK] button. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-19 9 Function Menu 9.3.7 Sample No. /ID Settings • Configure whether to display the sample number in the measurement results under [Edit and print No.]. • Configure whether to record the sequence number of the analyzer under [Store No.]. 9.3.8 Work List Selection Setting Configure whether to display the sample ID or patient ID at the Work list index screen. 9 9-20 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.9 Correlation Coefficients Enter, reset, or print the correlation coefficients for each sample type ([P/S,W] or [U]). For details, refer to “Description of the correlation function” at the end of the document. IMPORTANT: Incorrect entry of “a” and “b” causes incorrect measurement results. Always make sure that the entries (a, b) are correct. Next-page buttons Tap the [Data input] button. button is tapped, all correlation NOTE If the coefficient settings are printed. NOTE If the [Reset all tests] button is tapped, a reset confirmation screen appears. To reset the correlation coefficients of all measurement items, tap the [OK] button. To cancel the reset, tap the [Cancel] button. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-21 9 Function Menu Tap the button of the measurement item you want to edit. NOTE Only measurement items for which QC cards have been read are displayed. If the [Input] field of an item is tapped, the software keyboard (refer to Section 2.3) appears so you can enter the value. To change the sample type, tap the [P/S], [W], or [U] button. NOTE If the [Reset] button is tapped, a reset confirmation screen appears. To reset the displayed correlation coefficients for the sample type of the selected measurement item, tap the [OK] button. To cancel the reset, tap the [Cancel] button. NOTE If values are entered for "a" and "b" when values have already been entered, the current values are not discarded, and corrections are also applied. For details, refer to following “Description of the correlation function”. 9 9-22 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu IMPORTANT: Description of the correlation function This function is designed to determine the correlation between the measured data obtained using the FDC NX600 and the data obtained using the conventional measuring method with your own instruments. Y Y=aX+b FDC NX600 Conventional measuring method using your instruments Y (mg/dl) X Y=X Y=0.98X+5 600 The X-axis shows the measured data obtained using your instruments, and the Y- axis shows the measured data obtained using the FDC NX600. The correlation regression equation in this case is: Y = aX + b After values for the two coefficients (a, b) are registered in the analyzer memory, the FDC NX600 performs compensation calculations internally using the formula: X = (Y-b)/a. In this way, measured data obtained using the FDC NX600 are adjusted to match measured data that would be obtained using your instruments. Method In order to obtain a better correlation, it is essential to exercise caution regarding the amount of data and the sample. 1. FDC reference 2. 3. 0 0 A 10 600 B X (mg/dl) Measurement using your instruments 897N203029A FDC NX600 Instruction Manual 2022.01 You can configure a and b values by sample type. The correlation function is independent of other compensation functions. If values for (a, b) have already been input, old values will not be canceled as follows. If the old values are represented as (a1, b1), and the new values as (a2, b2), the resulting values for (a, b) will be determined according to the following formulas: a=a1×a2 b=a1×b2+b1 9-23 9 9 Function Menu 4. Depending on the values input for (a, b), the apparent limits of the measurement range will shift as follows. (Example) GLU measurements If a=1 and b=0 (Y=X), the measurement range is 10 to 600 mg/dl. But by inputting regression coefficients as shown, the lower limit changes to A and the upper limit to B. Printout example Value under the measurement range If the regression formula is Y = 0.98X + 5: A = 5 mg/dl B = 607 mg/dl Value above the measurement range Y (mg/dl) Y=1.1X-5 Y=X 600 If the regression formula is Y = 1.1X - 5: A = 14 mg/dl B = 550 mg/dl FDC reference 9 0 0 10 A B 600 X (mg/dl) Measurement using your instruments Printout example Value under the measurement range Value above the measurement range 9-24 If the measured data is below the measurement range, an “@” mark is added as shown in the example printout. If the measured data is above the measurement range, a “>” sign is printed as shown in the example printout. IMPORTANT: Reported values with “@” marks may not be accurate. Run the measurement again. 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.10 Lot Compensation Coefficients Enter the lot compensation coefficients (c, d, and e) printed on the QC cards of slides. This mode is needed if it is not possible to read the data directly because the QC card was lost or spoiled. IMPORTANT: Incorrect inputs for (c, d, and e) will cause incorrect measurement results. Make sure that the inputs for (c, d, and e) are correct using this mode. IMPORTANT: When the slide type number was changed, even if the new values (c, d, and e) are entered by using this setting, the new standard curve information will not be updated. It is necessary to read the QC card when the slide type number was changed. The slide type number is the most significant digit of a lot number. (Example) The slide type number of “Lot No.123456” is 1. Tap the button of the “Measurement item - sample type” you want to edit. NOTE Only “Measurement item - sample type” items that have been read from QC cards are displayed as buttons. NOTE Immuno-chemical items are not displayed. NOTE Measurement items are displayed in ascending order of test code, and sample type is displayed in the following order: [PS] and [W]. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-25 9 Function Menu Tap the button of the Lot No. for which you want to enter a lot compensation coefficient. button is tapped, all lot compenNOTE If the sation coefficient settings are printed. If the [Input] field of an item is tapped, the software keyboard (refer to Section 2.3) appears so you can enter the value. NOTE The most significant digit of a [Lot] number (the type number) is fixed. Enter from the second digit. NOTE Incorrect entries will cause incorrect measurement results. Always make sure that the displayed values are correct. button is tapped, all lot compenNOTE If the sation coefficient settings are printed. 9 NOTE For the expiration date for use (EXP), enter xxxx-xx (four digits for the year and two digits for the month). Also enter the hyphen (-). 9-26 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.11 Standard Curve Coefficients The calibration curve coefficients (A, B, C, D, E, and F) of the QC card information can be entered. IMPORTANT: Incorrect entry of calibration curve coefficients causes incorrect measurement results. Always make sure that the entered coefficients are correct. NOTE The calibration curve coefficients of this setting are entered automatically by QC card reading and calibration. <Configuring the lot number> 1. Tap the button of the measurement item for which the calibration curve is to be edited. NOTE Only CRP items for which QC cards have been read are displayed. 2. From the list display area, tap the lot number for which the calibration curve is to be edited. List display area 897N203029A FDC NX600 Instruction Manual 2022.01 9-27 9 9 Function Menu 3. If the [Input] field of an item is tapped, the software keyboard appears so you can enter the value. NOTE If the button is tapped, the calibration curve coefficients are printed. 4. First tap the [Input] field of [Lot]. The Lot number entry screen (software keyboard screen) appears. 5. Enter a lot number, and then tap the [OK] button. NOTE The most significant digit (the type number) is fixed. Enter from the second digit. The Calibration curve coefficient (A:) entry screen (software keyboard screen) appears. 9 9-28 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu <Configuration of calibration curve coefficients (A, B, C, D, E, and F)> 1. Enter the calibration curve coefficient (A:), and then tap the [OK] button. The Calibration curve coefficient (B:) entry screen (software keyboard screen) appears. 2. Enter B through F using the same procedure. <Configuring the expiration date for use> For the expiration date for use (EXP), enter xxxxxx (four digits for the year and two digits for the month). NOTE Also enter the hyphen (-). 9 IMPORTANT: Incorrect entries will cause incorrect measurement results. Always make sure that the displayed values are correct. button is tapped, values of the calibration curve coefficient configuration reNOTE When the sults are printed. 897N203029A FDC NX600 Instruction Manual 2022.01 9-29 9 Function Menu 9.3.12 Lot Qualification This function button appears (enabled) when USM function is enabled. It is possible to (dis-)qualify slide LOTs which are registered in analyzer. 1. Select a slide name which needs to change the setting. 2. (Dis-)qualification for selected slide LOT can be changed by selecting [ON] or [OFF] button. NOTE When the disqualified slide lot is measured, the test result will be indicated as “CHK QC” on screen, print and in communication. 9 9-30 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.13 PF Calibration Coefficients Configure the PF calculation correction coefficients. Next-page buttons Tap the [Data input] button. button is tapped, all PF NOTE I f t h e calculation correction coefficients are printed. NOTE If the [Reset all tests] button is tapped, a reset confirmation screen appears. To reset the PF calculation correction coefficients of all measurement items, tap the [OK] button. To cancel the reset, tap the [Cancel] button. Tap the button of the measurement item you want to edit. NOTE Only measurement items for which QC cards have been read are displayed as buttons. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-31 9 Function Menu If the [Input] field of an item is tapped, the software keyboard (refer to Section 2.3) appears so you can enter the value. NOTE If the [Reset] button is tapped, a reset confirmation screen appears. To reset the PF calculation correction coefficients of selected items, tap the [OK] button. To cancel the reset, tap the [Cancel] button. NOTE To cancel the configuration, tap the button. 9 9-32 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.14 Dilution Setting You can configure the dilution factor and position of the diluent to be used for each measurement item and sample type (PS and U). When a dilution factor is not configured for each sample on the sample information registration screen, measurement is performed using the dilution factor configured here. Next-page buttons 1. Tap the [Data input] button. NOTE When the button is tapped, all dilution factor settings will be printed. 2. Tap the button of the measurement item to be edited. 9 897N203029A FDC NX600 Instruction Manual 2022.01 9-33 9 Function Menu 3. For the dilution factor, tap a button, and then select from [None], [2], [3], [4], [5], or [10]. NOTE In the case of CRP, the selection menu to the left appears. 9 9-34 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.15 Tube Setting for Diluent & Reference Fluid Configure the test tube type to be used for diluent and reference fluid. NOTE For the types of containers that can be used for diluent and reference fluid, refer to Section 5.4 and Section 5.5. Tap the button of the container you want to use. 9.3.16 Print Setting for Measured Result Configure the print quantity (1 to 3 sheets) and print order of measurement results. • Tap the print quantity you want to configure. • For [Print order], you can select [Slide stacking order] or [Fixing]. NOTE If [Fixing] is selected, printing occurs in the following order. <Plasma/serum and urine> TP → ALB → TBIL → DBIL → GOT/AST → G P T / A LT → A L P → L D H → C H E → GGT → LAP → CPK → AMYL → TCHO → HDLC → TG → UA → BUN → CRE → Na-K-Cl → Ca → IP → Mg → GLU → CRP → NH3 → CKMB → LIP → TCO2 <Whole blood> Na-K-Cl → GLU → NH3 • For [Print with the ranges], to print out the ranges along with test results, check the checkbox. NOTE By default, the checkbox is checked (enabled). 897N203029A FDC NX600 Instruction Manual 2022.01 9-35 9 9 Function Menu 9.3.17 Control Fluid Setting Configure the control liquid name / lot number and acceptable range. 1. Tap the button of the control liquid to be edited. NOTE [-] (control liquid not registered) of Step 1 cannot be edited. 2. Tap a [Lot] button to choose a lot. 3. If you touch a display section (lower limit or upper limit) of the acceptable range to be edited, a software keyboard (Refer to Section 2.3) appears so you can edit the value. IMPORTANT: If the entry of the acceptable range is incorrect, the judgement of the control measurement results is incorrect. Always make sure that the entered values are correct. 9 NOTE The units configured by the [Unit conversion] of the function menu are displayed. NOTE If the acceptable range is changed, information from before the change is not saved. 4. To edit a lot number, touch the [Lot editing] button, and then enter the number using the software keyboard that appears. NOTE To edit or add a control liquid name, tap the button of a control liquid name, and then enter the information using the software keyboard that appears. 9-36 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.18 Control Measurement Setting Configure whether to enable the correlation coefficient when performing a control measurement. To enable the correlation coefficient for a control measurement, tap the checkbox to select it. NOTE By default, the checkbox is cleared (not enabled). 9.3.19 Operator ID Setting NOTE When the USM function is disabled, the following setting will appear. In case that the USM is enabled, refer to Section 11.3. Configure whether to display the Operator ID at the Measurement-preparation screen and measurement results screen. To display the operator ID, tap the checkbox. NOTE By default, the checkbox is cleared (not enabled). 897N203029A FDC NX600 Instruction Manual 2022.01 9-37 9 9 Function Menu 9.3.20 Test Item Name Setting You can configure the display of item names. The names appear in the screen display and printed paper of measurement results. 1 2 3 1 Choose the item names to be used for display and printing. 2 Choose the names to be used for external communication output to the host computer. The selection of [ALPi, LDHi]/[ALP, LDH] is the display setting of the test name ALPi, LDHi. This is not the display setting of the test name ALP, LDH. (ALPi and LDHi are slides compatible with the IFCC method. Availability depends on the region of sale.) NOTE If [ALP, LDH] is selected, ALPi, LDHi appears the same as ALP, LDH. Check this setting again before testing. 3 Choose the item names to be used. Tap either the [BUN] or [UREA] button. NOTE When the name of BUN is changed to UREA, or vice versa, set again the following items. ITEMS Reference interval Correlation coefficient Control fluid settings 9 NOTE When BUN/UREA is changed, the change is reflected in the following items: • Test Result • Unit conversion • Reference Interval setting • Calculated Parameter • Dilution factor Coefficients NOTE Regarding the BUN ↔ UREA conversion coefficient UREA = BUN×2.14 [mg/dL] UREA = BUN [mmol/L] NOTE Rebooting the device is necessary for the setting to take effect. The confirmation dialog for is tapped to exit Mode function for administrator. After the the power-off will appear when shutdown, reboot the analyzer. 9-38 897N203029A FDC NX600 Instruction Manual 2022.01 9 Function Menu 9.3.21 Default Sample Type Setting Set the default (initial setting) sample type. Tap a sample type to select it. 9.3.22 Password Setting When tapping the [Mode function for administrator] button on the Function menu screen, the password input may be required. The password can be changed here. If the input field is tapped, the software keyboard (refer to Section 2.3) appears so you can edit the password. NOTE A password can be 4 or 5 alphanumeric characters long. NOTE In the event of a password being lost, please contact your dealer for assistance. 897N203029A FDC NX600 Instruction Manual 2022.01 9-39 9 10 Other Functions 10.1 Data Communication The analyzer can transmit measurement results to a host computer or PC that has been approved by IEC/UL60950-1 or IEC/UL62368-1. To communicate, it is necessary to prepare the software and a cable to receive data. Before you use this function for the first time, please contact your dealer for assistance. NOTE Use a shielded LAN cable. Otherwise, electromagnetic interference may occur. NOTE When a LAN or RS-232C cable is not connected, measurement results are not sent. Make sure a cable is connected. NOTE When connecting this analyzer to a network, make sure it is placed inside of the firewall to avoid cybersecurity risks. 10 897N203029A FDC NX600 Instruction Manual 2022.01 10-1 11 USM (User Safety Mode) 11.1 Operator ID Input When Measurement Start and In Mode Function When measurement start button or function menu button is tapped, an Operator ID input display is shown. Then, input Operator ID. If input operator ID is registered and also within expiry, the measurement starts or the function menu becomes active. NOTE When not registered ID/expired ID/Emergency ID counted up to the maximum usage limit are input, measurement does not start, nor mode function screen become active. 11 897N203029A FDC NX600 Instruction Manual 2022.01 11-1 11 USM (User Safety Mode) 11.2 Difference of Operable Functions among Operator ID Level (OPE,ADM,EMG) Operable functions are different depending on operator ID level. NOTE OPE: Operator ADM: Administrator EMG: Emergency Measurement Result OPE/ADM: When expired slide is used, EXP is indicated instead of measurement result. EXP is used not only on screen indication, but also on print and in communication. This is also the case for CRP measurement with expired calibration. EMG: When expired slide is used in EMG, measurement result with # mark is indicated. 11 11-2 897N203029A FDC NX600 Instruction Manual 2022.01 11 USM (User Safety Mode) Mode function OPE/EMG: Operable function is limited to [Control], [Calibration], [START Manual pipetting] and some maintenance functions. 11 897N203029A FDC NX600 Instruction Manual 2022.01 11-3 11 USM (User Safety Mode) ADM: All functions are operable. 11 11-4 897N203029A FDC NX600 Instruction Manual 2022.01 11 USM (User Safety Mode) 11.3 USM Menu 11.3.1 Operator ID -Import 1. Create Operator ID data and save it by use of Operator ID Editor. * File name: dcn_uid_yyyymmddhhmmss.dat Ex) dcn_uid_20220425201612.dat * For the detail, see the manual of the Operator ID Editor. 2. Create the following folder structure in your USB. Rename the file which is created by the tool to [dcn_uid.dat], Store it in the USB, WData └ GUI └ dcn_uid.dat *Please use FUJIFILM recommended USB only. 3. Select [Operator ID setting] in ADM mode. 4. Select [Operator ID import]. 11 897N203029A FDC NX600 Instruction Manual 2022.01 11-5 11 USM (User Safety Mode) 5. The left pop-up is shown. Then set USB, and after a few seconds, select [OK]. 6. The left pop-up will be shown when import is complete. Take off the USB and then tap [Shutdown]. 11 11-6 897N203029A FDC NX600 Instruction Manual 2022.01 11 USM (User Safety Mode) 11.3.2 Operator ID -Export 1. Select [Operator ID setting] in ADM mode. 2. Select [Operator ID export]. 3. The left pop-up is shown. Then set USB, and after a few seconds, select [OK]. 11 897N203029A FDC NX600 Instruction Manual 2022.01 11-7 11 USM (User Safety Mode) 4. The left pop-up will be shown when export is complete. Then tap [OK], and take off the USB. 5. Operator ID data, [dcn_uid.dat], is stored in below folder structure below in the USB. WData └ GUI └ dcn_uid.dat * Browsing and editing of dcn_uid.dat is possible in Operator ID Editor. For the detail, see the manual of the Operator ID Editor. 11 11-8 897N203029A FDC NX600 Instruction Manual 2022.01 11 USM (User Safety Mode) 11.3.3 Editing Registered Operator ID Registered Operator ID can be edited as the following procedure. NOTE Editable items are the expiry of ID and the level. 1. Select [Operator ID setting] in Mode function for administrator screen. 2. Select [Operator ID setting]. 3. Select an Operator ID which needs editing. 11 897N203029A FDC NX600 Instruction Manual 2022.01 11-9 11 USM (User Safety Mode) 4. To change the expiry, touch the expiry field and edit it using the software keyboard. 5. To change the level, select either [ADM/ OPE]. 11 11-10 897N203029A FDC NX600 Instruction Manual 2022.01 11 USM (User Safety Mode) 11.3.4 Editing Emergency ID (Dis-)qualification of registered Emergency ID can be set. Also measurement limit count by use of Emergency ID can be set. 1. Select [Operator ID setting] in Mode function for administrator screen. 2. Select [Edit Emergency ID]. 3. To (dis-)qualify the Emergency ID, tap the checkbox. 11 897N203029A FDC NX600 Instruction Manual 2022.01 11-11 11 USM (User Safety Mode) 4. To edit the total limit count, select [Edit emergency ID count]. 5. To reset the current count, tap [Reset] button. To set a new limit count, select either [0], [50] or [100]. In case of 0, emergency ID does not function. 11 NOTE Measurement count by use of Emergency ID cannot exceed the limit count. When the current count is up to the limit, the measurement by use of Emergency ID is not possible. In this case, reset the current count. 11-12 897N203029A FDC NX600 Instruction Manual 2022.01 12 Specifications Processing speed 128 tests per hour (simultaneous CM and ISE measurements) Number of incubator cells CM: 13; ISE: 1 Incubation temperature 37°C (CM) / 30°C (ISE) Incubation time 1 to 6 minutes Capacity for simultaneously loaded slides 20 Slide lot information Each item: maximum 5 lots Sampler (1) Sample application fluid volume 50 µl or less (automatic switching) (2) Dilution Automatic dilution (maximum 21x dilution) (3) PF filtering time Max. 90 sec. Slide ejection Automatic ejection (max. 100 slides into the disposal box) Used tip ejection Automatic ejection (max. 50 tips into the disposal box) Measurement wavelengths 400 nm, 415 nm, 505 nm, 540 nm, 577 nm, 600 nm, 625 nm, 650 nm Light source Halogen lamp (6 V, 10 W) Measurement precision σ< = 0.0004 OD / 5 minutes Measurement accuracy Within ±0.005 OD (600 nm) Display 7-inch color touch panel Backup data of test results For 270 samples Printer Thermal type (paper size: 58 mm x 25 m) Warming up time Approx. 10 minutes / 25°C, approx. 20 minutes / 15°C Environmental conditions Indoor use (6,000 cd/m2 (lux) or lower. 3,000 cd/m2 (lux) or lower when using the sample barcode reader.) Altitude: Up to 2,000m (810hPa) Transient overvoltage category: II Pollution degree: 2 Ingress protection rating: IP00 (No Protection) Operating temperature: 15 to 32°C Operating humidity: 30 to 80% RH (no vapor condensation) Storage and transport conditions Temperature: -10 to 50°C Humidity: 10 to 90% RH (no vapor condensation) Voltage used Single phase AC; 100 - 240 V ±10%; 50 to 60 Hz Consumption current 1.9 - 0.8A Type of protection against electri- CLASS 1 EQUIPMENT cal shock Volume level Within 70 dB (including buzzer sound) Data transfer RS-232C (1 port); USB (2 port); LAN (1 port) External dimensions (W) 470× (D) 360× (H) 370 mm Weight FUJI DRI-CHEM NX600: 28Kg FUJI DRI-CHEM NX600i: 26Kg 897N203029A FDC NX600 Instruction Manual 2022.01 12 12-1 12 Specifications Effective period of service (service life) 6 years after installation (This period is valid as long as the precautions for use are followed and regular periodic maintenance is performed correctly.) 12 12-2 897N203029A FDC NX600 Instruction Manual 2022.01 13 Glossary Glossary of the display & printout messages NOTE Because the display spaces and printouts are limited, some abbreviations are used. NOTE Periods will not be used after abbreviations on the display and printout messages. Abbreviation Meaning AD Analog/Digital ADM administrator ave average Bef comp Before compress cali calibration charac character comm communication ctrl control DA Digital/Analog Decomp Decompress DL FUJI DRI-CHEM DILUENT DL (CRP) EMG emergency ERR error EXP expiration info information NG No Good OD Optical Density OPE operator Prob problem USM User Safety Mode 13 897N203029A FDC NX600 Instruction Manual 2022.01 13-1 26-30, NISHIAZABU 2-CHOME, MINATO-KU, TOKYO 106-8620, JAPAN https://www.fujifilm.com/ Heesenstrasse 31, 40549 Duesseldorf, Germany EU Oudenstaart 1, 5047 TK Tilburg, The Netherlands