PV 101+/PV 102+ Service Manual TABLE OF CONTENTS TABLE OF CONTENTS 1 INTRODUCTION............................................................................. 1–1 1.1 About the PV 101+ and PV 102+ ............................................................1–1 1.1.1 1.1.2 1.1.3 1.1.4 1.2 About this manual ...................................................................................1–2 1.2.1 1.2.2 1.2.3 2 Intended use of the PV 101+ ................................................................ 1–1 Intended use of the PV 102+ ................................................................ 1–1 Design and function .............................................................................. 1–1 Service personnel’s training requirements ............................................ 1–1 Scope .................................................................................................... 1–2 Intended audience................................................................................. 1–2 Icons...................................................................................................... 1–2 MAINTENANCE SERVICE INSTRUCTIONS ................................. 2–1 2.1 Purpose of the maintenance ...................................................................2–1 2.2 Introduction .............................................................................................2–2 2.2.1 2.2.2 2.2.3 2.2.4 2.2.5 2.3 Preparing inspection................................................................................2–4 2.3.1 2.3.2 2.3.3 2.3.4 2.3.5 2.4 Verifying the components and the installed software............................ 2–4 Initial recording...................................................................................... 2–4 Markings................................................................................................ 2–4 Information from the patient/user .......................................................... 2–4 Validity of the documentation ................................................................ 2–4 External inspection ..................................................................................2–5 2.4.1 2.4.2 2.4.3 2.5 Service check schedule ........................................................................ 2–2 Safety precautions ................................................................................ 2–2 Service record ....................................................................................... 2–2 Inspection equipment and tools ............................................................ 2–3 Replacement parts ................................................................................ 2–3 Visual inspection for external damage and wear .................................. 2–5 Checking the power supply connection................................................. 2–5 Short function check.............................................................................. 2–5 Internal inspection ...................................................................................2–6 2.5.1 2.5.2 2.5.3 2.5.4 2.5.5 2.5.6 2.5.7 2.5.8 2.5.9 2.5.10 Doc. No. 003259 En Opening the casing ............................................................................... 2–6 Cleaning the inside of the machine ....................................................... 2–6 Inspecting the internal wiring................................................................. 2–6 Securing the components ..................................................................... 2–6 Checking the power supply ................................................................... 2–6 Checking the electrical safety values .................................................... 2–6 Calibrating the pressure sensor ............................................................ 2–6 Calibrating the flow sensor.................................................................... 2–6 Calibrating the high pressure limit......................................................... 2–6 Reassembling the casing ...................................................................... 2–6 Issue: X-1 BREAS MEDICAL TOC – 1 TABLE OF CONTENTS 2.6 Final inspection before delivery .............................................................. 2–7 2.6.1 2.6.2 2.6.3 2.6.4 2.6.5 2.6.6 2.6.7 3 4 5 Function check ...................................................................................... 2–7 Checking the low pressure leakage alarm (PV 102+)........................... 2–7 Checking the trigger function ................................................................ 2–7 Checking the pressure and rate ............................................................ 2–7 Replacing the filters............................................................................... 2–8 Checking the accessories ..................................................................... 2–8 Adjusting the settings for the patient ..................................................... 2–8 REMOVING/REPLACING THE MAIN COMPONENTS ................. 3–1 3.1 Tools ....................................................................................................... 3–1 3.2 Removing and reassembling the upper casing ....................................... 3–2 3.3 Removing and replacing the CPU board ................................................ 3–3 3.4 Removing and reassembling the main assembly ................................... 3–4 3.5 Removing and replacing the blower unit................................................. 3–6 3.6 Removing and replacing the power board .............................................. 3–7 3.7 Removing and replacing the MDA board ................................................ 3–8 CALIBRATION INSTRUCTIONS.................................................... 4–1 4.1 Requirements.......................................................................................... 4–1 4.2 Equipment............................................................................................... 4–1 4.3 Calibrating the pressure sensor .............................................................. 4–2 4.4 Verifying the pressure calibration............................................................ 4–3 4.5 Calibrating the flow sensor...................................................................... 4–4 4.6 Calibrating the high pressure limit........................................................... 4–5 PARTS DRAWINGS AND PART-NUMBER LISTS ....................... 5–1 5.1 Overview – Main components................................................................. 5–2 5.2 Parts drawings ........................................................................................ 5–4 5.2.1 5.2.2 5.2.3 5.3 Parts drawing – Upper and lower casing .............................................. 5–5 Parts drawing – Main assembly ............................................................ 5–6 Parts drawing – Blower unit and patient air outlet................................. 5–7 Part-number lists..................................................................................... 5–8 5.3.1 5.3.2 5.3.3 5.3.4 6 PV 101+/PV 102+ Service Manual Part-number list – Upper and lower casing ........................................... 5–8 Part-number list – Main assembly......................................................... 5–9 Part-number list – Blower unit and patient outlet .................................. 5–10 Part-number list – Miscellaneous .......................................................... 5–11 FUNCTIONAL DIAGRAMS ............................................................ 6–1 6.1 Pneumatic diagram ................................................................................. 6–1 6.2 Functional block diagram ........................................................................ 6–2 TOC – 2 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7 ELECTRONICS............................................................................... 7–1 7.1 Electronics – Main components ..............................................................7–2 7.2 Functional block diagram ........................................................................7–4 7.3 Internal wiring diagram ............................................................................7–5 7.4 Circuit board descriptions........................................................................7–6 7.4.1 7.4.2 7.4.3 Electrical safety measurements ..............................................................7–8 7.6 Circuit diagrams – Circuit boards ............................................................7–9 7.7 7.8 Circuit diagram – CPU board 1, rev 2 ................................................... 7–10 Circuit diagram – CPU board 2, rev 2 ................................................... 7–11 Circuit diagram – CPU board 3, rev 2 ................................................... 7–12 Circuit diagram – CPU board 4, rev 2 ................................................... 7–13 Circuit diagram – CPU board 5, rev 2 ................................................... 7–14 Circuit diagram – Power board 1, rev 3................................................. 7–15 Circuit diagram – Power board 2, rev 3................................................. 7–16 Circuit diagram – MDA board 1, rev 3................................................... 7–17 Circuit diagram – MDA board 2, rev 3................................................... 7–18 Circuit diagram – MDA board 3, rev 3................................................... 7–19 Component location drawings – Circuit boards.......................................7–20 7.7.1 7.7.2 7.7.3 7.7.4 Component location drawing – CPU board........................................... 7–21 Component location drawing – Power board, upper side ..................... 7–22 Component location drawing – Power board, under side ..................... 7–23 Component location drawing – MDA board .......................................... 7–24 Component lists – Circuit boards ............................................................7–25 7.8.1 7.8.2 7.8.3 Component list – CPU board ................................................................ 7–25 Component list – Power board.............................................................. 7–26 Component list – MDA board ................................................................ 7–29 FAULT TRACING ........................................................................... 8–1 8.1 9 CPU board ............................................................................................ 7–6 Power board.......................................................................................... 7–6 MDA board ............................................................................................ 7–7 7.5 7.6.1 7.6.2 7.6.3 7.6.4 7.6.5 7.6.6 7.6.7 7.6.8 7.6.9 7.6.10 8 TABLE OF CONTENTS Fault-tracing table....................................................................................8–1 APPENDICES ................................................................................. 9–1 9.1 Engineering change history – PV 101+...................................................9–1 9.2 Engineering change history – PV 102+...................................................9–1 Service Record for BREAS PV 101+/PV 102+ 9.3 Returning products to BREAS.................................................................9–5 Service Report for BREAS PV 101+/PV 102+ Doc. No. 003259 En Issue: X-1 BREAS MEDICAL TOC – 3 TABLE OF CONTENTS PV 101+/PV 102+ Service Manual 10 PV 101+ OPERATING MANUAL 11 PV 102+ OPERATING MANUAL TOC – 4 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 1 INTRODUCTION INTRODUCTION This chapter gives an overview of the PV 101+ and PV 102+ devices and the service manual. WARNING! This product must be: • subjected to regular service, maintenance and control and any applicable upgrades, in accordance with BREAS service instructions; • repaired and/or modified in accordance with BREAS service manuals, technical bulletins and any special service instructions, by service technicians that have been authorised after BREAS PV 101+/PV 102+ service training, or have an equivalent technical knowledge on medical respiratory devices. Deviation from these service instructions may lead to risk of personal injury! 1.1 About the PV 101+ and PV 102+ The PV 101+ and PV 102+ are intended for use in homes, while travelling or in institutions. They are bi-level devices, which means that the expiration pressure (EPAP) can be set lower than the inspiration pressure (IPAP). 1.1.1 Intended use of the PV 101+ The BREAS PV 101+ device is intended for treatment of obstructive sleep apnoea. 1.1.2 Intended use of the PV 102+ The BREAS PV 102+ device is intended for treatment of obstructive and central sleep apnoea, hypopnea and other forms of sleep-related breathing problems. The back-up rate is adjustable, as well as the rise time, which makes it possible to customise the time for the pressure to change from EPAP to IPAP. 1.1.3 Design and function The PV 101+ and PV 102+ are based on a blower which is driven by an electronically controlled servomotor. The microprocessor-based electronics calculates, from the set pressure, rate, rise time etc., the correct motor speed. The set pressure and trigger sensitivity is continuously monitored. An audible alarm is immediately given in the event of a power failure. 1.1.4 Service personnel’s training requirements Service personnel working with the PV 101+ and PV 102+ devices should have medical/technical training and a good knowledge of the construction and function of respiratory devices. Authorisation by BREAS PV 101+/PV 102+ service training is recommended. Always contact your BREAS representative if you have any questions or if training is required. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 1–1 INTRODUCTION PV 101+/PV 102+ Service Manual 1.2 About this manual 1.2.1 Scope All the maintenance checkpoints and the additional service actions for the PV 101+/ PV 102+ are described in this manual. It contains all documentation required for maintaining and servicing the machines, such as enlarged drawings, wiring diagrams, component location guides and so on. BREAS Medical reserves the right to make changes to the products and/or the contents of this manual without any prior notice. 1.2.2 Intended audience This service manual is intended for service technicians who have medical/technical training and who have a good knowledge of the construction and function of respiratory devices. The service manual is NOT intended for clinic personnel or patients, who can find all the information they need in the PV 101+ and PV 102+ Operating Manuals. 1.2.3 Icons Icons are used in this manual to highlight specific types of information. The meaning of each icon is explained in the table below. Icon Explanation Warning! Risk of death and serious personal injury. Caution! • Risk of minor or moderate injury. • Risk of equipment damage, loss of data, extra work or unexpected results. Note Information that may be valuable but is not of critical importance, tips. Reference Reference to other manuals with additional information on a specific topic. 1–2 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 2 MAINTENANCE SERVICE INSTRUCTIONS MAINTENANCE SERVICE INSTRUCTIONS This chapter describes all the routine checks and additional service instructions for the PV 101+ and PV 102+. WARNING! This product must be: • subjected to regular service, maintenance and control and any applicable upgrades, in accordance with BREAS service instructions; • repaired and/or modified in accordance with BREAS service manuals, technical bulletins and any special service instructions, by service technicians that have been authorised after BREAS PV 101+/PV 102+ service training, or have an equivalent technical knowledge on medical respiratory devices. Deviation from these service instructions may lead to risk of personal injury! For information about fault tracing, parts drawings, board schematics and so on, refer to the respective chapter in this service manual. The patient and/or care provider should follow the function checks described in the chapter “CHECKS BEFORE USE” in the PV 101+/PV 102+ Operating Manual. 2.1 Purpose of the maintenance The PV 101+ and PV 102+ are designed to give the users many years of trouble-free assistance, provided that preventive maintenance is carried out at the intervals specified in this manual. Well-performed maintenance will considerably increase the life of the machine. It is also important to check any peripheral equipment at the same time as services are carried out. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 2–1 MAINTENANCE SERVICE INSTRUCTIONS 2.2 PV 101+/PV 102+ Service Manual Introduction Before you start the maintenance service, read the safety precautions and make sure you have a new service record and all the necessary equipment, tools and replacement parts at hand. 2.2.1 Service check schedule The schedule below lists the recurrent service checks of the PV 101+/PV 102+’s components and the intervals at which they should be performed. The reference numbers refer to chapters and sections in this manual that contain more detailed instructions. A complete maintenance service, as described in this chapter, must be carried out at least every 12 months. Interval Service check See section Every 12 months Complete maintenance check 2 Check the electrical safety limit values. 7.5 Calibrate the pressure, flow and high pressure limit sensors. 4 Replace the blower unit. 3.5 Calibrate the pressure, flow and high pressure limit sensors. 4 Every 15,000 operating hours or when required 2.2.2 Safety precautions Follow the safety precautions below when working with the PV 101+/PV 102+: • Do not work on the device with the casing removed and any power supply connected unless the instructions in this manual clearly say so. • Always use caution when working with the device connected to the mains and the casing removed. • Make sure that all precautions to prevent electrostatic discharge (ESD) have been taken. Follow all the regulations regarding ESD. • Explosive gases and liquids must not be kept or used near the PV 101+/PV 102+. The PV 101+/PV 102+ Operating Manual contains an extended list of safety precautions. 2.2.3 Service record See section 9, “APPENDICES” for the BREAS service record. • 2–2 Copy the service record and use it for noting the service checks while performing the yearly service. BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 2.2.4 MAINTENANCE SERVICE INSTRUCTIONS Inspection equipment and tools Before servicing the PV 101+/PV 102+, make sure you have the following equipment at hand: • Test lung • Multimeter • Measuring equipment for pressure 0–70 mbar (Thommen HM 28 digital manometer is recommended, BREAS part no. 001934) • Flow meter • Screwdrivers TX10, TX25, Phillips 1 • Box socket wrenches M3, M4 • Allen key 2, 5x25 For the equipment required when calibrating the device, see section 4.2. 2.2.5 Replacement parts The following parts should be available when servicing the device: Description BREAS part no. Patient tube 000245 Patient air filter, washable (2 pcs) 000996 Patient air filter, disposable (if used) 001259 Air filter, internal cooling 001121 Blower unit assembly, 15.000 hours 003110 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 2–3 MAINTENANCE SERVICE INSTRUCTIONS PV 101+/PV 102+ Service Manual 2.3 Preparing inspection 2.3.1 Verifying the components and the installed software • Check section 9.1, “Engineering change history – PV 101+” or section 9.2, “Engineering change history – PV 102+” to find out what changes have been made to the device and from which serial number they were implemented. If in any doubt about the circuit board revisions or upgrades that may have been made that have not been recorded, check the component designations on the circuit boards. 2.3.2 2.3.3 2.3.4 2.3.5 Initial recording 1 Copy a new service record (see section 9, “APPENDICES”). 2 Identify the PV 101+/PV 102+; note the model and serial number and any inventory number, if applicable, on the service record. 3 Check earlier actions and events recorded on any previous service records. 4 Note the current patient settings. 5 Note the total number of operating hours. Markings Check that the following markings are legible: • Make, model designation, serial number • Warning texts • Any inventory markings • Any other texts on the external equipment and accessories Information from the patient/user Before starting the service, check the following with the patient/clinic: • Has the PV 101+/PV 102+ functioned without any problems? If not, what were they? • How does the patient/care provider check the function of the device? How often? • How often are the filters changed? • How many filters will be required before the next service? • Other observations? Validity of the documentation 1 Check that the documentation used by the patient is valid and up to date. 2 2–4 Check if any modification or revision of the PV 101+/PV 102+ needs to be done at the same time as the service. BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual MAINTENANCE SERVICE INSTRUCTIONS 2.4 External inspection 2.4.1 Visual inspection for external damage and wear 1 Clean the cover of the PV 101+/PV 102+ using a mild detergent. 2.4.2 2.4.3 2 Check for visible damage to the cover and other components. 3 Check that all external parts are tightened and secured. Checking the power supply connection 1 Check the plug, power cord and mains power socket. 2 Make sure the strain-relief clamp for the power cord is not damaged. 3 Check the power supply to the machine. 4 If the PV 101+ does not operate from the mains power, check the fuses. Short function check 1 Connect the power cord to the power socket of the PV 101+/PV 102+. 2 Connect the patient circuit. 3 Start the PV 101+/PV 102+ and check that it functions normally. 4 Disconnect the power cord and check that the power failure alarm sounds. Reconnect the power cord. The alarm should be silenced. 5 Switch on the PV 101+/PV 102+. Upon starting, all the LEDs on the front panel should be lit and the trigger settings should be indicated on the bar graph. The entire bar graph should then become yellow for 2 seconds before the device starts functioning normally. 6 (PV 102+ only) Switch on the PV 102+ and wait for 15 seconds for the leakage alarm to be activated. 7 Switch off the device. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 2–5 MAINTENANCE SERVICE INSTRUCTIONS 2.5 PV 101+/PV 102+ Service Manual Internal inspection Disconnect the power supply before removing the casing of the machine. 2.5.1 Opening the casing • For instructions, see section 3.2, “Removing and reassembling the upper casing”. 2.5.2 Cleaning the inside of the machine • Remove any dust or dirt that has collected in the machine. 2.5.3 Inspecting the internal wiring • Inspect all wiring and the connectors for damage. Make sure that no wires or cables are pinched. 2.5.4 Securing the components • Make sure that all the internal components, such as the motor unit, the circuit boards, the connectors and so on, are properly secured. 2.5.5 Checking the power supply 1 Check that the power inlet socket is undamaged and that it is properly fastened. 2 Make sure the power board is properly fastened. 3 Check the wiring to and from the power board. 2.5.6 Checking the electrical safety values • For instructions, see section 7.5, “Electrical safety measurements”. 2.5.7 Calibrating the pressure sensor • For instructions, see section 4.3, “Calibrating the pressure sensor”. 2.5.8 Calibrating the flow sensor • For instructions, see section 4.5, “Calibrating the flow sensor”. 2.5.9 Calibrating the high pressure limit • For instructions, see section 4.6, “Calibrating the high pressure limit”. 2.5.10 Reassembling the casing • For instructions, see section 3.2, “Removing and reassembling the upper casing”. 2–6 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual MAINTENANCE SERVICE INSTRUCTIONS 2.6 Final inspection before delivery 2.6.1 Function check 1 Connect the power cord to the power socket of the PV 101+/PV 102+. 2.6.2 2 Connect the patient circuit. 3 Start the PV 101+/PV 102+ and check that it functions normally. 4 Disconnect the power cord and check that the power failure alarm sounds. Reconnect the power cord. The alarm should be silenced. 5 Switch on the PV 101+/PV 102+. Upon starting, all the LEDs on the front panel should be lit and the trigger settings should be indicated on the bar graph. The entire bar graph should then become yellow for 2 seconds before the device starts functioning normally. 6 (PV 102+ only) Switch on the PV 102+ and wait for 15 seconds for the leakage alarm to be activated. 7 Switch off the device. Checking the low pressure leakage alarm (PV 102+) 1 Adjust the pressure settings to: 2 2.6.3 Checking the trigger function 1 Adjust the pressure settings to: 2 3 2.6.4 IPAP 10 mbar EPAP 10 mbar Make sure that the alarm LED is lit and that an audible alarm is heard. IPAP 10 mbar EPAP 4 mbar Connect the patient tube without a mask. Block the air outlet on the tube. When you remove the blockage, a triggered breath should be initiated. Checking the pressure and rate Pressure The measurement should be made using a blocked patient circuit with leakage provided (see figure below). 1 2 Adjust the pressure settings to: IPAP 10 mbar EPAP 10 mbar Measure the pressure and check that it is correct (tolerance: ±4% of the maximum value and ±8% of the current pressure value). Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 2–7 MAINTENANCE SERVICE INSTRUCTIONS PV 101+/PV 102+ Service Manual Pressure gauge mbar 5 mm leakage Patient circuit Plug Rate 3 Adjust the settings to: IPAP 10 mbar EPAP 4 mbar Back-up rate 8 BPM (PV 102+ only) The PV 101+ back-up rate is fixed to 6 BPM. 4 2.6.5 2.6.6 Replacing the filters The patient air filters are located in the filter holder under the carrying handle of the device. The internal cooling filter is located in the lower casing. 1 Replace the washable grey air filter. 2 Replace the disposable white filter, if used. 3 Replace the internal cooling filter from underneath the machine. Checking the accessories 1 Inspect the patient circuit. Replace it if necessary. 2 2.6.7 2–8 Measure the rate and check that it is correct. Inspect any other accessories. Adjusting the settings for the patient • Adjust all the patient settings to the prescribed values. BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 3 REMOVING/REPLACING THE MAIN COMPONENTS REMOVING/REPLACING THE MAIN COMPONENTS This chapter describes how to install, remove and/or replace the main components of the PV 101+/PV 102+. 3.1 Tools Before disassembling the PV 101+/PV 102+, make sure you have the following tools at hand: • Screwdriver TX10 • Screwdriver TX25 • Screwdriver Phillips 1 • Box socket wrench M4 • Box socket wrench M3 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 3–1 REMOVING/REPLACING THE MAIN COMPONENTS 3.2 PV 101+/PV 102+ Service Manual Removing and reassembling the upper casing Disconnect the power supply before removing the casing of the machine. 1 2 To remove the casing: Remove the mains power cord. Turn the PV 101+/PV 102+ upside down. Remove the four screws placed in the four holes underneath the lower casing and the two screws placed underneath the handle. 6 × TX25 3 Lift up the upper casing. Disconnect the CPU board’s ribbon cable from the MDA board. Put the upper casing aside. To reassemble the casing: • Reassemble in reverse order. 3–2 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 3.3 REMOVING/REPLACING THE MAIN COMPONENTS Removing and replacing the CPU board To remove the CPU board: 1 Remove the upper casing (See section 3.2). 2 Turn it upside down and remove the three Torx screws placed underneath the CPU board. 3 Remove the CPU board. 3 × TX10 To reassemble the CPU board: • Reassemble in reverse order. Fit the components to the holes for the display panel controls and screw the CPU board back in place. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 3–3 REMOVING/REPLACING THE MAIN COMPONENTS 3.4 PV 101+/PV 102+ Service Manual Removing and reassembling the main assembly All the main internal components – blower unit, power board and MDA board – are mounted on the base plate. To remove the main assembly: 1 Remove the upper casing (See section 3.2). 2 Turn the PV 101+/PV 102+ upside down. Remove the M4 nut that fastens the blower unit rubber mount. 3 Remove the five screws on the lower casing that hold the base plate. 2 1 × M4 Compound nut 3 5 × TX10 3–4 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 4 REMOVING/REPLACING THE MAIN COMPONENTS Press the rear panel side of the lower casing outwards to release the connectors and lift out the main assembly. To reassemble the main assembly: • Reassemble in reverse order. Make sure the connectors on the rear panel are properly fitted in their holes. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 3–5 REMOVING/REPLACING THE MAIN COMPONENTS 3.5 PV 101+/PV 102+ Service Manual Removing and replacing the blower unit To remove the blower unit: 1 Remove the upper casing (See section 3.2). 2 Remove the main assembly (See section 3.4). 3 Disconnect the motor connectors from the MDA board. 4 Disconnect the flex tube from the blower unit’s air outlet. Disconnect the red sensor tube from the blower unit’s sensor connector. 5 Remove the remaining two nuts that hold the blower unit’s rubber mounts from underneath the base plate. 6 Lift away the complete blower unit. 3 3 5 4 2 × M4 Compound nut 5 To reassemble the blower unit: • 3–6 Reassemble in reverse order. BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 3.6 REMOVING/REPLACING THE MAIN COMPONENTS Removing and replacing the power board To remove the power board: 1 Remove the upper casing (See section 3.2). 2 Remove the main assembly (See section 3.4). 3 Remove the blower assembly (See section 3.5). 4 Move aside the patient air and sensor tubing. Disconnect the power cables to the external DC and mains supplies on the MDA board. 5 Disconnect the DC control cable from the PSU board. 6 Unscrew the two screws that hold the rear part of the power board, and remove the two nuts under the base plate. 7 Pinch the two securing pins on the front part of the power board, e.g. by using a screwdriver, and lift up the power board from the base plate. 2 × Phillips 1 5 4 6 7 To reassemble the power board: • Reassemble in reverse order. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 3–7 REMOVING/REPLACING THE MAIN COMPONENTS 3.7 PV 101+/PV 102+ Service Manual Removing and replacing the MDA board To remove the MDA board: 1 Remove the upper casing (See section 3.2). 2 Remove the main assembly (See section 3.4). 3 Remove the blower assembly (See section 3.5). 4 Remove the power board (See section 3.6). 5 Disconnect the patient air and sensor tubing from the sensor connectors on the MDA board. 6 Disconnect the external DC cable, the fan cable and the DC control cable from the MDA board. 7 Turn the base plate upside down and remove the 3 screws that fix the MDA board to the bottom of the plate. 8 Remove the 4 screws that fix the MDA board to the rear part of the base plate. 9 Lift up the MDA board from the base plate. 4 × T10 8 4 × M3 Compound nut 6 5 7 3 × T10 3 × M3 Compound nut To reassemble the MDA board: • 3–8 Reassemble in reverse order. BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 4 CALIBRATION INSTRUCTIONS CALIBRATION INSTRUCTIONS This chapter describes how to adjust and verify the pressure, flow and high pressure limit sensors of the PV 101+/PV 102+ device. 4.1 Requirements • The calibration should be made every 12 months or after 15,000 operating hours. • Keep the power cord of the PV 101+/PV 102+ connected to the mains power during the calibration. Always use caution when working with the device connected to the mains and the casing removed. Always check and adjust the patient settings to the prescribed values after you have calibrated the PV 101+/PV 102+. 4.2 Equipment Make sure you have the following equipment to hand: • Screwdriver for the calibration screws (∅ 1 mm) Pressure calibration • Flow restrictor nipple 6 mm, BREAS part number 000997 • Flow restrictor nipple 5 mm, BREAS part number 003311 • Voltmeter • Syringe with a y-piece tube • Pressure gauge, 0–70 mbar Flow calibration • Flow restrictor 6 mm, BREAS part number 000997 • Voltmeter High pressure limit calibration • Flow restrictor nipple 4 mm, BREAS part number 003312 • Fan control unit, BREAS part number 003302 • Pressure gauge, 0–70 mbar Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 4–1 CALIBRATION INSTRUCTIONS 4.3 PV 101+/PV 102+ Service Manual Calibrating the pressure sensor 1 Start the PV 101+/PV 102+ device by pressing the I/O button and the + button at the same time. 2 Remove the casing (see section 3.2). Keep the CPU board connected to the MDA board. 3 Use the voltmeter to measure the voltage on pin 9 on the analogue/digital output connector on the rear panel of the PV 101+/PV 102+. 4 Adjust the screw on the P203 (Offset) potentiometer on the MDA board until the voltage is 1V. 5 Connect the pressure gauge to the syringe and the y-piece tube. Remove the black tube from the pressure sensor on the MDA board. Connect the free end of the y-piece tube to the pressure sensor. 6 Use the syringe to create a pressure of 30 mbar ±0.5 mbar. 7 Adjust the screw on the P204 (Gain) potentiometer on the MDA board until the voltage is 4 V. 8 Remove the y-piece tube from the MDA board’s pressure sensor and reconnect the black tube. 9 Repeat step 4 to 8 until no further adjustment is necessary. 10 Remove the equipment. Continue with the pressure verification procedure (see section 4.4). P203 P204 Pressure sensor MDA board P203 Offset BREAS MEDICAL P204 Gain _ 4–2 _ + + Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 4.4 CALIBRATION INSTRUCTIONS Verifying the pressure calibration 1 Start the PV 101+/PV 102+. 2 Remove the casing (see section 3.2). 3 Connect a pressure gauge to the patient air inlet and arrange a leakage with a 5 mm flow restrictor (see diagram below). 4 Set both the IPAP and EPAP pressure to 20 mbar. 5 Check that the pressure rises to 20 mbar ±1 mbar. If not, the P204 potentiometer must be adjusted. 6 Set both the IPAP and EPAP pressure to 4 mbar. 7 Check that the pressure rises to 4 mbar ±0.5 mbar. If not, the P203 potentiometer must be re-adjusted. 8 If necessary, repeat the pressure calibration procedure (see section 4.3). 9 Remove the equipment and reassemble the casing (see section 3.2). Pressure gauge mbar 5 mm leakage Patient circuit Doc. No. 003259 En Issue: X-1 BREAS MEDICAL Plug 4–3 CALIBRATION INSTRUCTIONS 4.5 PV 101+/PV 102+ Service Manual Calibrating the flow sensor Always calibrate the pressure sensor before you calibrate the flow sensor. Calibrating the flow sensor first may lead to unreliable results. 1 Start the PV 101+/PV 102+ device by pressing the I/O button. 2 Plug the patient air outlet with a 6 mm flow restrictor nipple. 3 Set both the IPAP and EPAP pressure to 12 mbar. 4 Remove the casing (see section 3.2). Keep the CPU board connected to the MDA board. 5 Use the voltmeter to measure the voltage on pin 8 on the analogue/digital output connector on the rear panel of the PV 101+/PV 102+. 6 Adjust the screw on the P201 (Gain) potentiometer on the MDA board until the voltage is 2.15 V (±0.15 V). 7 Block the air outlet to completely stop the air flow. 8 Adjust the screw on the P202 (Offset) potentiometer on the MDA board until the voltage is 0.5 V (±0.05 V). 9 Remove the air flow blockage. 10 Repeat step 6 to 9 until no further adjustment is necessary. 11 Remove the equipment and reassemble the casing. P201 P202 MDA board P201 Gain BREAS MEDICAL P202 Offset _ 4–4 + + _ Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 4.6 CALIBRATION INSTRUCTIONS Calibrating the high pressure limit 1 Start the PV 101+/PV 102+ by pressing the I/O button. 2 Connect a pressure gauge to the patient air inlet and arrange a leakage with a 4 mm flow restrictor nipple (see figure). Pressure gauge mbar 4 mm leakage Patient circuit Plug 3 Remove the casing (see section 3.2). Disconnect the CPU board from the MDA board. 4 Connect the manual fan control unit to the X202 connector on the MDA board. 5 Use the fan control unit to start the PV 101+/PV 102+ blower at full speed. 6 Adjust the screw on the P205 potentiometer on the MDA board until the pressure measured on the pressure gauge is 40 mbar (±1 mbar). 7 Remove the equipment and reassemble the casing. P205 X202 MDA board + Doc. No. 003259 En Issue: X-1 _ P205 BREAS MEDICAL 4–5 CALIBRATION INSTRUCTIONS 4–6 PV 101+/PV 102+ Service Manual BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 5 PARTS DRAWINGS AND PART-NUMBER LISTS PARTS DRAWINGS AND PART-NUMBER LISTS This chapter contains enlarged drawings and part-number lists of the PV 101+/PV 102+. The main components of the device are displayed at the beginning of the chapter. Some of the listed parts can be ordered as replacement parts from BREAS. Please contact your BREAS supplier for more information. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 5–1 PARTS DRAWINGS AND PART-NUMBER LISTS 5.1 PV 101+/PV 102+ Service Manual Overview – Main components 2 1 3 4 5 6 7 8 5–2 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual PARTS DRAWINGS AND PART-NUMBER LISTS No Description 1 Upper casing 2 Push-button overlay On the upper casing 3 CPU board In the upper casing 4 Blower unit, incl. motor Mounted on base plate 5 Power board Mounted to bottom of base plate 6 MDA board Mounted to rear of base plate 7 Base plate In the lower casing 8 Lower casing Doc. No. 003259 En Issue: X-1 Location BREAS MEDICAL 5–3 PARTS DRAWINGS AND PART-NUMBER LISTS 5.2 5–4 PV 101+/PV 102+ Service Manual Parts drawings BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 5.2.1 PARTS DRAWINGS AND PART-NUMBER LISTS Parts drawing – Upper and lower casing For definitions of the part numbers, refer to the list in section 5.3.1. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 5–5 PARTS DRAWINGS AND PART-NUMBER LISTS 5.2.2 PV 101+/PV 102+ Service Manual Parts drawing – Main assembly For definitions of the part numbers, refer to the list in section 5.3.2. 5–6 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 5.2.3 PARTS DRAWINGS AND PART-NUMBER LISTS Parts drawing – Blower unit and patient air outlet For definitions of the part numbers, refer to the list in section 5.3.3. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 5–7 PARTS DRAWINGS AND PART-NUMBER LISTS 5.3 Part-number lists 5.3.1 Part-number list – Upper and lower casing 5–8 PV 101+/PV 102+ Service Manual BREAS Part no. Definition 000041 Rubber mount 000488 Absorbent, dwg 10646-1 (T2), outer edge upper casing push-buttons 000489 Absorbent, dwg 10646-2 (T2), outer edge inlet bend 000490 Absorbent, dwg 10646-3 (T2), outer edge inlet 000492 Absorbent, dwg 10646-5 (T2), outer edge outer wall 000494 Absorbent, dwg 10646-7 (T2), outer edge upper casing inlet 000497 Absorbent, dwg 10647-3 (T2) outer edge lower casing outer wall 000499 Absorbent, dwg 10647-5 (T2) outer edge lower casing outer wall 000501 Absorbent, dwg 10647-7 (T2) outer edge lower casing outer wall 000674 M4, wing nut 000678 M4, lock nut 000706 Power cord strain relief 000718 Screw, MRX steel M5 x 12 fzb POZ 000996 Filter PV 101+/PV 102+, washable (2 pcs) 001095 Outer casing, upper 001112 Button for CPU board 001121 Filter 45 x 45 x 10 for cooling fan 001245 Screw, RX-PT 4-20*6 fzb 001259 Filter, disposable (5 pcs) 002490 Screw Mrt M4x14 fzb 002931 Overlay, back panel PV 101+/PV 102+ 002960 CPU board PV 101+/PV 102+ 002962 Lower casing PV 101+/PV 102+ 002963 Screw Rts St 3.5 X 6.5, PV 101+/PV 102+ 003198 Overlay, PV 102+ setting panel, Swedish 003199 Overlay, PV 102+ setting panel, English 003200 Overlay, PV 102+ setting panel, German 003201 Overlay, PV 102+ setting panel, Finnish 003202 Overlay, PV 102+ setting panel, Spanish 003203 Overlay, PV 102+ setting panel, Italian 003204 Overlay, PV 102+ setting panel, Dutch BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 5.3.2 PARTS DRAWINGS AND PART-NUMBER LISTS BREAS Part no. Definition 003205 Overlay, PV 102+ setting panel, Danish 003206 Overlay, PV 102+ setting panel, French 003207 Overlay, PV 101+ setting panel, Swedish 003208 Overlay, PV 101+ setting panel, English 003209 Overlay, PV 101+ setting panel, German 003210 Overlay, PV 101+ setting panel, Finnish 003211 Overlay, PV 101+ setting panel, Spanish 003212 Overlay, PV 101+ setting panel, Italian 003213 Overlay, PV 101+ setting panel, Dutch 003214 Overlay, PV 101+ setting panel, Danish 003215 Overlay, PV 101+ setting panel, French 003249 Decal, PV 101+ logo 003250 Decal, PV 102+ logo 003261 Sticker “Plus” Part-number list – Main assembly BREAS Part no. Definition 000071 Mains input socket 000529 Spacer cooling fan PV 101+/PV 102+ 000789 Bolt MRX-H-M4x12H fzb 001041 Fan (cooling air PV 101+/PV 102+) 002959 Power board PV 101+/PV 102+ 002961 MDA board PV 101+/PV 102+ 1310005 (Eribel part no) M3 Compound nut 1310011 (Eribel part no) DSS M3050x8 Spacer 1310012 (Eribel part no) 5 mm DSUB Spacer 1310020 (Eribel part no) MRXZ 3x6 Steel Screw 1310029 (Eribel part no) RXS-Z ST2, 9x9, 5 Screw 1310110 (Eribel part no) Spacer with securing pin, 10 mm 1310114 (Eribel part no) DSS M3050x10 Spacer 1360074 (Eribel part no) Base plate PV 101+/PV 102+ 1410006 Mains power fuse, 110–220V, 1A T 5x29 mm Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 5–9 PARTS DRAWINGS AND PART-NUMBER LISTS 5.3.3 5–10 PV 101+/PV 102+ Service Manual Part-number list – Blower unit and patient outlet BREAS Part no. Definition 000032 Pressure sensor - MPX 10 DP 000640 Y-connector 000675 M4, Serrated washer 000677 Nut, M6M M4 Metal fzb 000789 Bolt MRX-H-M4x12H fzb 001000 Blower cover, upper 001001 Blower cover, lower with silencer 001003 Blower fan wheel 001004 Bushing, blower fan wheel 001005 Patient air inlet 001027 O-ring 7.66 x 1.78 001028 O-ring 38 x 1 001032 LF isolator 001033 Honeycomb material 001038 Flex tube ∅ 22 x 15 mm 001039 Silicon tube, 5 x 8 mm (blue/green) 001052 Silicon tube, 50 mm (red) 001053 Silicon tube, 30 mm (red) 001054 Silicon tube, 125 mm (blue) 001055 Silicon tube, 150 mm (green) 001056 Silicon tube, 70 mm (black) 001074 Spacer DSSM4070x25 001079 Blower casing, upper 001080 Blower casing, lower 001081 Sealing cover for upper casing 001082 Sealing cover for upper blower casing 001083 Rubber bushing 001111 Blower casing outlet 22 mm 001117 Flow sensor AWM2100V 001122 Seal for upper blower casing 001245 Screw, RX-PT 4-20*6 fzb 001665 Shoulder screw M4 4,3 002966 Compound nut M4, PV 101+/PV 102+ BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 5.3.4 PARTS DRAWINGS AND PART-NUMBER LISTS BREAS Part no. Definition 003110 Turbine Exchange kit PV 101+/PV 102+ Part-number list – Miscellaneous BREAS Part no. Definition 000245 Hytrel patient tube 000491 Absorbent, dwg 10646-4 (T2) outer edge outer wall 000503 Absorbent, dwg 10648-1 (T2) inner edge upper casing outer wall 000504 Absorbent, dwg 10648-2 (T2) inner edge upper casing round 000505 Absorbent, dwg 10648-3 (T2) inner edge upper casing U 000506 Absorbent, dwg 10648-4 (T2) inner edge upper casing bottom 000507 Absorbent, dwg 10648-5 (T2) inner edge upper casing outer 000508 Absorbent, dwg 10648-6 (T2) inner edge upper casing outer 000509 Absorbent, dwg 10648-7 (T2) inner edge upper casing outer 000510 Absorbent, dwg 10648-8 (T2) inner edge upper casing outer 000511 Absorbent, dwg 10649-1 (T3) inner edge lower casing inner 000512 Absorbent, dwg 10649-2 (T3) inner edge lower casing inner 000522 Absorbent dwg 10665-1 silencer 1001 round 000523 Absorbent dwg 10665-2 silencer 1001 inside 000711 Cable tie 71 mm 000997 Flow restrictor nipple 6 mm 001073 Washer, 12 x 5 x 2 001934 Thommen HM 28 digital manometer 002965 Screw Mrt M5 X 12 Torx, PV 101+/PV 102+ 002973 PV 101+ Operating manual, Swedish 002974 PV 101+ Operating manual, German 002975 PV 101+ Operating manual, Finnish 002976 PV 101+ Operating manual, English 002977 PV 101+ Operating manual, Norwegian 002978 PV 101+ Operating manual, Spanish 002979 PV 101+ Operating manual, Italian 002980 PV 101+ Operating manual, Dutch 002981 PV 101+ Operating manual, Danish Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 5–11 PARTS DRAWINGS AND PART-NUMBER LISTS 5–12 BREAS Part no. Definition 002983 PV 101+ Operating manual, Portuguese 002982 PV 101+ Operating manual, French 002984 PV 101+ Operating manual, Greek 002985 PV 101+ Operating manual, Polish 002986 PV 102+ Operating manual, Swedish 002987 PV 102+ Operating manual, German 002988 PV 102+ Operating manual, Finnish 002989 PV 102+ Operating manual, English 002990 PV 102+ Operating manual, Norwegian 002991 PV 102+ Operating manual, Spanish 002992 PV 102+ Operating manual, Italian 002993 PV 102+ Operating manual, Dutch 002994 PV 102+ Operating manual, Danish 002995 PV 102+ Operating manual, French 002996 PV 102+ Operating manual, Portuguese 002997 PV 102+ Operating manual, Greek 002998 PV 102+ Operating manual, Polish 003036 EMC-protection, Flow meter 003116 Screw K6sf M4x16mm 003220 Connector 003221 Connector element 003243 PV 101+ Operating manual, Japanese 003244 PV 102+ Operating manual, Japanese 003259 PV 101+/PV 102+ Service manual, English 003302 Fan control unit 003311 Flow restrictor nipple 5 mm 003312 Flow restrictor nipple 4 mm BREAS MEDICAL PV 101+/PV 102+ Service Manual Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 6 FUNCTIONAL DIAGRAMS FUNCTIONAL DIAGRAMS This chapter contains a diagram of the pneumatic system of the ventilator and a block diagram of the PV 101+/PV 102+’s functions. 6.1 Pneumatic diagram The pneumatic diagram shows the pneumatic components of the air circulation system of the PV 101+/PV 102+. Sensor 2 Red Sensor 1 Green Blue Black No. Description 1 Patient air inlet (through air filter) 2 Blower 3 Motor 4 Silencer 5 Flow sensor (2) 6 Pressure sensor (1) 7 Honeycomb material 8 Reference pressure tube 9 Patient air tube Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 6–1 FUNCTIONAL DIAGRAMS 6.2 PV 101+/PV 102+ Service Manual Functional block diagram The block diagram below shows how the electronics are designed and how they are connected to the other components. 6–2 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7 ELECTRONICS ELECTRONICS This chapter describes in detail the electronics of the PV 101+/PV 102+. It also contains circuit diagrams and component location drawings for the circuit boards of the machine. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–1 ELECTRONICS 7.1 PV 101+/PV 102+ Service Manual Electronics – Main components The electronics in the PV 101+ and the PV 102+ comprises a blower unit and three printed circuit boards: the CPU board, the power board and the MDA board. 1 2 3 4 5 6 8 7–2 BREAS MEDICAL 7 Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual ELECTRONICS No Description Location 1 CPU board In the upper casing 2 Blower unit, incl. motor Mounted on base plate 3 Power board Mounted to bottom of base plate 4 MDA board Mounted to rear of base plate 5 Hour counter Mounted to MDA board 6 Analogue/digital connection Mounted to MDA board 7 Battery connection Mounted to rear of MDA board 8 Power connection Mounted to rear of base plate Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–3 ELECTRONICS 7.2 PV 101+/PV 102+ Service Manual Functional block diagram The block diagram below shows on an overall level how the electronics are designed and how they are connected to the other components. 7–4 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7.3 ELECTRONICS Internal wiring diagram The wiring diagram displays the internal cabling and connections of the electronics. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–5 ELECTRONICS 7.4 PV 101+/PV 102+ Service Manual Circuit board descriptions This section contains detailed descriptions and figures of the PV 101+/PV 102+’s circuit boards. 7.4.1 CPU board All the push-buttons are mounted on the CPU board together with the LEDs and the three bar-graphs for pressure display. Also mounted here is the four 7-segment displays for the rise time and backup rate displays (PV 102+). The microprocessor with its software is also located here, as well as an internal memory (DALLAS) for the logging and storage of the patient usage data. Refer also to the circuit diagrams (See section 7.6.1, 7.6.2, 7.6.3, 7.6.4 and 7.6.5) and component location drawings (See section 7.7.1). 7.4.2 Power board The power supply unit is fed primarily with mains power via the mains power inlet to the internal AC/DC converter. Secondly, the power board can be fed with 12–24 DC via the external battery input. Refer also to the circuit diagrams (See section 7.6.6 and 7.6.7) and component location drawings (See section 7.7.2 and 7.7.3). The power supply has mains power as soon as the power cord is connected. 7–6 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7.4.3 ELECTRONICS MDA board X201 connector P203 P204 P201 P202 X202 connector Flow sensor Pressure sensor The MDA (motor drive assembly) board comprises the driver stage for the blower motor, the pressure sensor, the flow sensor, the hour counter and the D-shell connector for serial communication. The driver stage is built around the U205 control circuit and the Q205–Q210 transistors. The P21 buzzer is powered by the C221 supercap if the mains power supply should fail. The pressure sensor, which is of the semi-conductor type, is amplified by U201. The offset and gain can be adjusted using the P200 and P201 potentiometers. The signal is used for regulating the pressure. The flow sensor is also of the semi-conductor type, a mass flow meter. The flow that passes the sensor is proportional to the flow through the flow resistance in the air tube from the blower. The offset and gain can be adjusted using the P202 and P203 trim potentiometers. The signal is used for initiating the trigger and the mask-off function. Refer also to the circuit diagrams (See section 7.6.8, 7.6.9 and 7.6.10) and component location drawings (See section 7.7.4). Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–7 ELECTRONICS 7.5 PV 101+/PV 102+ Service Manual Electrical safety measurements Electrical safety measurements must be made in accordance with EN 60 601-1. However, you can make an insulation resistance measurement instead of the voltage test specified by the standard. The measurements can be made using an automatic electrical safety tester. All tests must be performed in accordance with class II type BF. Supply voltage • Note the power voltage reading. The voltage must be noted for each service check, as the currents measured are directly in relation to the supply voltage. This allows all measurements made on the same machine to be compared with measurements made on different occasions. Insulation • Measure the insulation resistance using a 500V DC power supply. The most suitable method is to connect the plus lead to the two ventilator power socket pins and the minus lead to the casing or patient air connector. The measurements made during the delivery inspection can be used as reference values for measurements made during future services. If no reference values are available, the value for the insulation resistance should be >20MΩ. Leakage currents • Measure the leakage currents at different parts of the ventilator using an RC circuit to ground. The measurements are made partly at normal case (NC) and at the single fault condition (SFC). Reverse the polarity of the power supply and note the highest value. The leakage currents to ground must not exceed the stated limit values. Leakage currents from the casings • Measure the leakage current of the casing at an unpainted point, for example, the head of a screw. Limit values: NC SFC Break neutral for SFC. < 0.1mA < 0.5mA Patient leakage currents • Measure the patient leakage currents between the patient connector and ground. Limit values: NC SFC Break neutral for SFC. < 0.1mA < 0.5mA Leakage currents with mains power supply at the patient connected part • Perform this test using an automatic electrical safety tester with this function. See the safety instructions for the tester. Limit value: 7–8 SFC < 5mA BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7.6 ELECTRONICS Circuit diagrams – Circuit boards Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–9 C109 100n 100n C134 C114 100n 100n C135 C115 100n 100n C138 C128 T6 T5 RES P54 P55 C100 100n C101 100n VCC 1 3 4 5 8 13 14 7 R102 10k Vcc V- 16 2 6 15 9 12 11 10 VCC P23 C102 100n C103 100n F1 C104 P57 A12 A13 A14 22pF VCC P56 A11 P21 P55 A15 P54 A10 A9 A8 A7 A6 P63 P62 P61 P60 A0 A1 A2 A3 A4 A5 P64 P65 P66 P67 1 26 35 52 53 54 55 56 57 58 59 60 61 62 63 64 27 28 29 30 31 32 33 34 D7 D6 D5 D4 D3 D2 D1 D0 XTAL100 VCC P60 P61 P62 P63 P64 P65 P66 P67 P50 P51 P52 P53 P54 P55 P56 P57 P20 P21 P22 P23 P24 P25 P26 P27 10MHz U102 NMI Vcc Vss BA LIR R/W WR RD E Vss XTAL EXTAL MP0 MP1 RES STBY A15 A14 A13 A12 A11 A10 A9 A8 A7 A6 A5 A4 A3 A2 A1 A0 HD63C03YRF 36 37 38 39 40 41 42 43 A7 A6 A5 A4 A3 A2 A1 A0 D0 A14 A13 A8 A9 A11 /OE A10 /CS D7 44 45 46 47 48 49 50 51 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 U104 D7 A10 A11 A9 A8 A13 A14 29F010 29 28 27 26 25 24 23 22 21 D7 D6 D5 D4 D3 D2 D1 D0 P21 P23 P24 P25 P50 P51 P52 C105 22pF A0 A1 A2 A14 A15 A0 A1 A2 A3 A13 P53 1k P54 R107 P55 P56 P57 P60 P61 P62 P63 P64 P65 P66 P67 A14 A13 A12 A11 A10 A9 A8 A7 A6 A5 A4 A3 A2 A1 A0 1 26 2 23 21 24 25 3 4 5 6 7 8 9 10 2 3 4 5 6 7 8 9 1 11 12 13 U106:D 19 18 17 16 15 14 13 12 10 D2 D3 D4 D5 D6 D7 D7 D6 D5 D4 D3 D2 D1 D0 11 D1 4 5 1 2 9 10 D0 PV101/102 Upgrade U106:B 74HC00 U106:A U106:C 74HC00 6 3 8 Date 74HC00 Sheet 19 18 17 16 15 13 12 11 14 20 27 22 74HC00 I/O7 I/O6 I/O6 I/O5 I/O4 I/O3 I/O2 I/O0 U105 I1 I2 I3 I4 I5 I6 I7 I8 GND D7 D6 D5 D4 D3 D2 D1 D0 GND CE WE OE CLK OE GAL16V8_PLCC U103 A14 A13 A12 A11 A10 A9 A8 A7 A6 A5 A4 A3 A2 A1 A0 DS1644 Title Axel Olofsson CPU/Display cad056a_R3.sch Number Drawn by A3 Size Filename CE2 CE3 CE4 CE9 CE8 CE7 CE6 CE5 D0 D1 D2 D3 D4 D5 D6 D7 Rev of 2002-10-30 2 5 Issue: X-1 Doc. No. 003259 En U101 C1+ GND V+ C2+ C1- C2R2OUT R1OUT T1IN T2IN DS14C232 R2IN R1IN T1OUT T2OUT P24 VCC R103 R104 R105 R106 4*4k7 1 2 3 4 5 6 7 8 9 10 3,3k M1 MODUL_1 VCC A7 A6 A5 A4 A3 A2 5 6 7 8 9 10 11 12 13 D3 P53 D0 D4 A1 A0 D5 P52 D6 U100 8 7 6 5 D1 Vcc WP SCL SDA A12 4 3 2 1 32 31 30 A12 A15 A16 NC1 Vcc /WE NC2 D1 D2 GND D3 D4 D5 D6 14 15 16 17 18 19 20 D2 A0 A1 A2 Vss 24LC01B-I 3,3k P56 1 2 3 4 R101 P57 10u/6V 10u/6V P25 + C111 100n 100n P50 100n C129 P51 T1 100n C139 VCC 100n C133 VCC 100n + C140 + C110 C108 C132 BREAS MEDICAL 7–10 Circuit diagram – CPU board 1, rev 2 7.6.1 PV 101+/PV 102+ Service Manual ELECTRONICS C131 100n 10u/6V R100 1 11 U109 10 1 2 3 4 5 6 7 8 9 10 M2 S2 D7 LD2 LD100 D6 D7 D6 D7 CE6 CE8 RES D7 D6 D5 D4 D3 D2 D1 D0 D7 D6 D5 D4 D3 D2 D1 D0 D0 D1 D2 D3 D4 D5 D6 D7 D5 D5 CE9 D1 D2 D3 D4 D1 D2 D3 D4 D0 D0 CE2 3 4 7 8 13 14 17 18 1 11 3 4 7 8 13 14 17 18 1 11 2 3 4 5 6 7 8 9 1 11 2 3 4 5 6 7 8 9 74HC273 1D 2D 3D 4D 5D 6D 7D 8D CLR CLK U118 74HC273 1D 2D 3D 4D 5D 6D 7D 8D CLR CLK U114 74HC574 1D 2D 3D 4D 5D 6D 7D 8D OC CLK U115 74HC574 1D 2D 3D 4D 5D 6D 7D 8D OC CLK 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8D GND 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8D GND 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q GND 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q GND LD107 470R R117 LD105 2 5 6 9 12 15 16 19 10 2 5 6 9 12 15 16 19 10 19 18 17 16 15 14 13 12 10 19 18 17 16 15 14 13 12 DA_S1_8 DA_S1_7 DA_S1_6 DA_S1_5 DA_S1_4 DA_S1_3 DA_S1_2 DA_S1_1 DA_S2_8 DA_S2_7 DA_S2_6 DA_S2_5 DA_S2_4 DA_S2_3 DA_S2_2 DA_S2_1 16Q 15Q 14Q 13Q 12Q 11Q 10Q 9Q 470R R118 LD106 470R M3 R-2R (10K-20K) 1 2 3 4 5 6 7 8 9 10 S1 R-2R (10K-20K) D6 POWER ON/OFF OK100 - KEY OK101 FUNCTION OK102 SLEEP OK104 + KEY OK103 D4 D5 D6 D7 D0 D1 D2 D3 D0 D1 D2 D3 D4 D5 3 4 7 8 13 14 17 18 1 11 2 3 4 5 6 7 8 9 1 11 1D 2D 3D 4D 5D 6D 7D 8D OC CLK U116 OK1 P56 P57 P55 P54 74HC273 1D 2D 3D 4D 5D 6D 7D 8D CLR CLK U117 74HC574 560R R128 470R R114 LD102 1k R109 GND 100k R111 470R R110 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8D GND 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q 2 5 6 9 12 15 16 19 10 19 18 17 16 15 14 13 12 10 DP4 DP3 DP2 DP1 LD103 L119-EGW_2 470R R113 LD101 470R 470R R115 R108 R116 1Q 8Q 7Q 6Q 5Q 4Q 3Q 2Q P51 R127 2k2 P25 P51 P50 T1 T6 T5 1 2 5 6 8 9 12 13 10u/6V C112 + 4093 GND Vcc U112 D5 D6 Q104 7 14 3 4 10 11 16 15 14 13 12 11 10 9 + U119 R R WDO N.C N.C WDI N.C GND BC847B D3 D2 D1 D0 D7 D6 D5 D4 D3 D2 D1 D0 D7 CE7 CE5 1 2 3 4 5 6 7 8 4k7 R112 C136 100n C113 10u/6V MAX699CWE GND Vcc Vcc GND N.C N.C N.C N.C R123 GND Vcc F G A B C D E GND Vcc F G A B C D E GND Vcc F G A B C D E GND Vcc F G A B C D E 47R 4511 LT BI LE A B C D 4511 U111 LT BI LE A B C D 4511 U110 LT BI LE A B C D 4511 U108 LT BI LE R126 2k2 3 4 5 7 1 2 6 3 4 5 7 1 2 6 3 4 5 7 1 2 6 3 4 5 U107 A B C D Q103 VCC 1 2 3 4 RN103 RN105 2k2 6 3 5 15 14 13 7 4 1 2 3 4 4027 J Vcc CLK Q K Q S R J CLK K Q S Q R GND U113 1 2 3 4 8 7 6 5 RN102 1 2 3 4 RN104 RN106 16 1 2 12 11 10 9 8 C117 1n 8 7 6 5 1k C120 1n T6 C121 1n BC847B Q101 Q100 BC847B R119 C119 100n 47R C118 1n 2k2 R124 Q102 8 7 6 5 8 7 6 5 8 7 6 5 R121 MNR34-J5AJ-471_2 8 7 6 5 MNR34-J5AJ-471_2 1 2 3 4 MNR34-J5AJ-471_2 8 7 6 5 8 7 6 5 BC847B C116 1n 47R R122 RN107 MNR34-J5AJ-471_2 1 2 3 4 RN100 MNR34-J5AJ-471_2 MNR34-J5AJ-471_2 1 2 3 4 1 2 3 4 RN101 MNR34-J5AJ-471_2 MNR34-J5AJ-471_2 R125 VCC VCC VCC VCC BC847B 8 16 15 14 13 12 11 10 9 8 16 15 14 13 12 11 10 9 8 16 15 14 13 12 11 10 9 8 16 15 14 13 12 11 10 9 T1 7 1 2 6 P50 D4 P51 LD104 VCC 5 10 9 7 6 4 2 1 5 10 9 7 6 4 2 1 5 10 9 7 6 4 2 1 5 10 9 7 6 4 2 1 C126 100n C122 100n S2 CE4 C130 R132 C127 100n C123 1n 47R R120 3 8 DP4 3 8 DP3 3 8 DP2 3 8 DP1 LD2 VCC Filename Drawn by A3 Size 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 1ar 2ar 3ar 4ar 5ar 6ar 7ar 8ar 9ar 10ar 1ag 2ag 3ag 4ag 5ag 6ag 7ag 8ag 9ag 10ag 1ar 2ar 3ar 4ar 5ar 6ar 7ar 8ar 9ar 10ar 1ag 2ag 3ag 4ag 5ag 6ag 7ag 8ag 9ag 10ag 1ar 2ar 3ar 4ar 5ar 6ar 7ar 8ar 9ar 10ar 1ag 2ag 3ag 4ag 5ag 6ag 7ag 8ag 9ag 10ag OK1 S1 S2 T6 T5 P25 P51 P50 T1 LD2 VCC P54 P55 P57 P56 21 22 23 24 25 26 27 28 29 30 21 22 23 24 25 26 27 28 29 30 21 22 23 24 25 26 27 28 29 30 cad056a_R3.sch Axel Olofsson CPU/Display Sheet Date PV101/102 Upgrade 3Q 4Q 5Q 6Q 7Q 8Q 1Q 2Q 3Q 4Q LED106 MV59164 5Q 6Q 7Q 8Q 1Q 2Q 3Q 4Q 5Q 6Q LED105 MV59164 7Q 8Q 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q LED104 MV59164 Number C124 C125 C107 100n 100n 100n Title TDSG1160 LED103 470R TDSG1160 R131 LED102 470R R130 TDSG1160 470R LED101 R129 TDSG1160 LED100 470R S1 CE3 100n 1k R133 R134 P25 C137 T5 BREAS MEDICAL 100n OK1 20 19 18 17 16 1k 2k 3k 4k 5k 10k 9k 8k 7k 6k 11 12 13 14 15 20 19 18 17 16 1k 2k 3k 4k 5k 10k 9k 8k 7k 6k 11 12 13 14 15 20 19 18 17 16 1k 2k 3k 4k 5k 10k 9k 8k 7k 6k Doc. No. 003259 En Issue: X-1 11 12 13 14 15 16Q 15Q X100:8 Rev X100:3 X100:2 X100:13 X100:12 X100:11 X100:9 X100:7 X100:6 X100:5 X100:4 X100:10 X100:1 X100:14 2 of 2002-10-30 11Q 12Q 13Q 14Q 15Q 16Q 9Q 10Q 11Q 12Q 13Q 14Q 15Q 16Q 9Q 10Q 11Q 12Q 13Q 14Q 15Q 16Q 9Q 10Q 11Q 12Q 13Q 14Q 2 5 7.6.2 470k PV 101+/PV 102+ Service Manual ELECTRONICS Circuit diagram – CPU board 2, rev 2 7–11 R145 R144 10k 10k 10k 10k 10k 10k 10k R135 R146 20k R147 R136 20k R137 20k R138 20k R139 20k R140 20k R141 20k R142 20k R143 20k R148 F1 P60 P61 P62 P63 P64 P65 P66 P67 R149 BI1 BI2 BI3 BI4 BI5 BI6 BI7 BI8 BI9 BI10 T1 R150 Title Size PV101/102 Upgrade Date 2 2002-10-30 Rev Sheet CPU/Display Axel Olofsson Number cad056a_R3.sch A3 Drawn by of Filename 2 5 Issue: X-1 Doc. No. 003259 En BREAS MEDICAL 7–12 Circuit diagram – CPU board 3, rev 2 7.6.3 PV 101+/PV 102+ Service Manual ELECTRONICS BI10 BI9 BI8 BI7 BI6 BI5 BI4 BI3 BI2 S2 DA_S2_8 DA_S2_7 DA_S2_6 DA_S2_5 DA_S2_4 DA_S2_3 DA_S2_2 DA_S2_1 20k R159 20k R158 20k R157 20k R156 20k R155 20k R154 20k R153 20k R152 20k R160 R161 R162 R163 R164 R165 BREAS MEDICAL R166 R151 10k 10k 10k 10k 10k Doc. No. 003259 En Issue: X-1 10k cad056a_R3.sch Filename CPU/Display Sheet Date PV101/102 Upgrade Axel Olofsson Number Drawn by A3 Size Title 2 of 2002-10-30 Rev 2 5 7.6.4 10k BI1 PV 101+/PV 102+ Service Manual ELECTRONICS Circuit diagram – CPU board 4, rev 2 7–13 DA_S1_1 DA_S1_2 DA_S1_3 DA_S1_4 DA_S1_5 DA_S1_6 DA_S1_7 DA_S1_8 S1 R178 R177 R176 10k 10k 10k 10k 10k 10k 10k R167 R179 20k R180 R168 20k R169 20k R170 20k R171 20k R172 20k R173 20k R174 20k R175 20k R181 BI1 BI2 BI3 BI4 BI5 BI6 BI7 BI8 BI9 BI10 R182 Title Size PV101/102 Upgrade Date 2 2002-10-30 Rev Sheet CPU/Display Axel Olofsson Number cad056a_R3.sch of Drawn by A3 Filename 2 5 Issue: X-1 Doc. No. 003259 En BREAS MEDICAL 7–14 Circuit diagram – CPU board 5, rev 2 7.6.5 PV 101+/PV 102+ Service Manual ELECTRONICS CN2:3 CN2:2 CN2:1 100-240V AC 4k7 6 5 2 3 - R3 1 U1:B 4k7 R4 PWR_CTRL 7 LM324Y U1:A CTRL_IN CTRL_OUT LM324Y 8k2 + - + Ledig SG37 R6 9 10 - + 100n 2 R1 C59 R61 100n 470k C54 L1 220n 2n2 Y1 C3 8 4k7 R9 +5V LM324Y U1:C 11 220uH C5 12 13 U1:D 4 15k D3 +VB 14 2k2 R59 +VB LM324Y R16 + - R15 680R 10uF/16V + LM324Y GND Vcc U1:E 100n C2 2n2 Y1 C4 TR1 +VB TR4 R17 BZX84C5V1 CNY17F-3 U2 470R 2 3 1 8 7 6 5 C6 OPTO_A 1 D1 R2 BZX84C3V6 3k3 C55 4k7 R8 100n R7 R5 D2 1M/0.6W 680R BZX84C5V1 C1 R10 R62 4k7 + C58 R14 R11 4k7 VARI_S14K300 R12 0u33/35V C9 CN1 C7 C8 OUT GND GND NC 4 6 5 78L05ACM U9 68nF IN GND GND NC BR1 1 2 3 4 - DB107G +5V + C57 1n/400V C13 + BC817-40 Q1 1k R22 Q2 BZX84C24 D4 BC817-40 Q3 BC807-40 R28 1 2 3 4 5 6 7 REF NC Vcc Vc OUT GND PGND 8k2 R31 UC 3845D COMP NC VF NC SENSEC NC RC U3 R32 D5 400V_DC 5k6/2W 14 13 12 11 10 9 8 DIOD_SMC 4n7/400V XY Monteras ej C12 100n 1n/400V 1n/400V R18 3k3 R19 18k 1n/400V 4k7 56R 100u/400V C15 R24 R25 R26 100k R13 1k5 1k5 C14 R23 R91 10n 3k9 47R 1n/400V R27 4k7 270k 47k 8k2 220k/0.6W R29 9 1 2 13 33541-901 Q4 100k R102 1k R34 150R R33 BC807-40 (9) 11 (7) (11) TR2 TRAFO_ETD34 BAS21 D8 D9 + 47k R99 100k SPP20N60S5 Q5 G BC807-40 Q16 100k R56 Q18 S D BC817-40 BYD37M D10 C22 330p/400V BYT12PI-1000 R98 C21 10u/50V Q17 R36 270R/0.6W 12 (10) 5 (3) 7 (5) 10 (8) AC_PWR BYD37M 10R C24 R42 C61 C16 220k 100n/400V R104 R105 C20 + R30 2n2 D21 R101 C18 1k5 68R C17 33u/50V 330p C66 BZX84C20 100k + 22k + R92 4n7/400V XY BAS21 C10 R20 R21 + C11 C26 1000uF/50V BC807-40 Q13 100k R53 L2 22uH +VB Filename TP3 + 4 8 - + 6 5 +5V 100k R52 DC_OUT 100n PSU - AC/DC Power cad055a_R3.sch CN2:6 CN2:7 CN2:4 CN2:5 03-01-23 1 of Sheet Rev Date Breas Medical AB LM393 GND Vcc U4:C 100n C28 BC817-40 Q6 D U4:B C53 LM393 7 Q7 "PMOS_TO220" RXE 185 R90 A. Olofsson Number +5V Drawn by A3 Size Title R60 CONNECTED TO PCB SHIELD R35 22u R97 D7 R38 R39 0R68/1W 100k 220p/400V C65 D13 100k R103 4k7 68k 100n D6 100n BYD37M C60 R37 BZX84C22 R100 33R R40 G + 0R68/1W 47k D12 C29 S R51 C25 47k D11 100n 6k8 K1 2k2 100k C23 R43 R54 R55 C27 R50 R48 R49 3 2 TP5 10k 1000uF/50V BZX84C10 6k8 10k 4k7 BREAS MEDICAL 4k7/2W 22k R57 R58 Doc. No. 003259 En Issue: X-1 10k NTC 7.6.6 47u/16V PV 101+/PV 102+ Service Manual ELECTRONICS Circuit diagram – Power board 1, rev 3 7–15 2n2 C19 S3 ISO ISO 1k R64 R78 Q10 Q15 R95 100k R96 MTD20P06HDL D15 U8 CNY17F-3 2k2 R84 1 3 BZX84C5V1 TR3 TRAFO_ETD29 D18 5 12 10 7 D19 BYD37M D22 BYT12PI-1000_2 D20 AC_PWR ? L3 CN2:8 Breas Medical AB PSU - AC/DC Power Date Sheet Bat/Main 2 of 03-01-23 3 100n Rev DC_OUT BZX84C6V8 D S 4 ISO Number A. Olofsson C52 BC807-40 100k G D S C56 cad055a_R3.sch 100u/50V 100k Q9 Q14 ISO Q12 STP60NE06-16FP_2 R77 10R G Title Size 100R 10k A3 100R R87 Drawn by R89 10k R75 C42 100n R88 1 3 2 Q8 BC817-40 ISO R79 D16 R93 10k ISO C44 OPTO_A 3 1 2 100k 680p U6 CNY17F-3 L4 R71 R74 + 100n C50 220p/400V Filename + C51 C35 R68 10k 10u/20V C47 C63 220p/400V R66 G BAS21 Q11 STP60NE06-16FP_2 10R 4n7/50V 47R/1W 47k 16 15 14 13 12 11 10 9 D17 R76 R83 C31 220p/400V C64 220p/400V U7 INV VREF NI VCC E/A OUTB CLK VC RT PGND CT OUTA RAMP GND SS ILIM/SD 1k R86 5 6 4 1k 0R1/1W CN3 47R D 1k R94 R72 R81 R82 R70 1 2 3 4 5 6 7 8 UC3825 C41 R80 C46 4n7/50V 47R/1W C45 R85 S 100n 6k8 10n ISO C43 100k BAS21 1 6.3A Trög C39 4k7 100n 68k R69 12k C40 680p 2n2 C36 10k + R65 C37 R67 100n 47u/6V 56R 100n 10n C38 2 12-24V DC 1000u/50V 5 4 6 1000u/50V C34 R63 + C33 + C32 2 Issue: X-1 Doc. No. 003259 En BREAS MEDICAL 7–16 Circuit diagram – Power board 2, rev 3 7.6.7 PV 101+/PV 102+ Service Manual ELECTRONICS BYW29-500 D14 R73 470k 0R1/1W 1 OUT+ GND OUT- Vcc SENSOR200 3 C271 LM274 GND Vcc U201:E 100nF C204 4 4 2 +10V TRYCKGIV MPX10DP 11 +5V X206:7 X206:1 X206:4 X206:5 X206:6 X206:2 X206:3 BAS32_2 X207 2 1 BAS32_2 D202 MOLEXBATTIN_VINK C201 C280 C202 2 P203 IN OUT 3,3K 1K 3 R255 R201 1 GND OFFSET {Value} U213 5k 3,3K R200 220uF/50V C220 3 2 6 5 + + - - D204 LM274 U201:A 1 LM274 7 U201:B {Value} RXE020 S200 BAS32_2 5 1 REL200 TIMER CURTIS701P GND IN 4 2 3 4 3 NTC 47k 100R R202 100R R214 P204 20k 6 2 4 2 4 8 3 GAIN BC847B Q215 IN OUT GND FB ON/OFF GND Q203 LM2597M-12 LM2576S-5 1 3 5 U210 IN OUT Vbs FLAG SD/SS GND DELAY FEEDBACK BC857B_2 Q200 7 1 5 6 D225 270R R203 270R R216 D215 330uH 9 10 10R R292 BYG90-40 BC847B Q216 BYG90-40 DR206 HM50-331K 330uH 10K R236 4.7k 8 3,3K R291 C221 LM274 R208 + D213 +5V 100uF/25V U201:C 4.7k R217 - +12V + BAS85_2 C244 DR209 HM50-331K + Monteras ej!!! R232 1K OUT OUT OUT OUT 2 7 3 6 BC847B Q201 TL317C IN NC NC ADJ U202 R205 0,47F 1 8 5 4 2 3 13 12 1 - + - + LM393 1 U206:A LM274 14 U201:D 2 3 + 3,3K R233 +10V R256 10R R270 1 8 5 4 680R 1K R230 2,2K R231 +5V BC847B OUT OUT OUT OUT TL317C IN NC NC ADJ U203 R213 Q202 PZ200 OBO-1206C DSS306-91FZ103N 1,6K R210 DR205 10K R235 BAS85_2 D212 R211 Vcc R 100nF D200 100nF Y1 100nF Y1 + R319 VR203 C281 220R 3,3K R234 + C209 1 2 Mont Ej +35V C206 RESET TIMER R218 R219 4,7uF/16V TIMER200 C212 X205:6 X205:7 C207 C203 1K D214 1 1 2 C210 DR208 BAS85_2 2 3 3 1 Filename 2 100nF Axel O cad057a_R3.sch Number MDA 1 Sheet of 2003-01-14 Rev X202:10 X202:9 X200:9 X200:6 X202:1 X202:7 X200:1 X202:2 X202:8 X202:3 X200:7 Date PV101/102 Upgrade MDA 1nF C216 1nF C217 3 DR210 DSS306-91FZ103N DSS306-91FZ103N Drawn by A3 Size Title D219 BAS85_2 220R 10uF/6V D203 DR201 DSS306-91FZ103N 2 7 3 6 R212 + 1,5K R204 R209 560R 4k7 150k 47k VR204 SM6T15CA X205:4 X205:5 100nF 100nF 39k 18k C245 C211 C205 C254 0,1uF/50V Mont Ej C213 +V 1nF 10k R207 R206 1k 10k R215 100nF C261 100nF C256 C259 100nF C260 100nF C257 R293 Mont Ej C258 100nF C255 D205 BREAS MEDICAL SM6T15CA Doc. No. 003259 En Issue: X-1 C253 100nF 100nF 100nF 3 3 7.6.8 Mont Ej PV 101+/PV 102+ Service Manual ELECTRONICS Circuit diagram – MDA board 1, rev 3 7–17 +12V + 1 SENSOR201 +10V + LM274 14 U214:D 5k1 R312 +10V R238 10k U214:A 1 LM274 R311 1k Q214 RDF16N06LESM 1k X202:5 5 6 - + LM393 7 U206:B 1K R229 R228 1 +5V 1 3 3 2 3 33uF/16V C237 DR202 DSS306-91FZ103N 2,2K DR204 DSS306-91FZ103N 2 2 DR203 DSS306-91FZ103N 1 +12V D208 D207 D206 SM6T15CA SM6T15CA DR207 BL01 RN1-A62 Title 100nF Number Axel O MDA X200:8 X202:11 X202:12 X200:5 X203:2 X203:1 X200:2 X200:3 1 2003-01-14 Rev Date of Sheet PV101/102 Upgrade MDA C264 cad057a_R3.sch 3 3 Issue: X-1 Doc. No. 003259 En + - 7 10k 10K R284 100K R283 - + R237 3 270R LM358Y 2 R225 270R R226 +10V P202 1k + - 7 4.7k U209:B R223 10K R282 4.7k R224 +12V 5 6 C238 0,1uF/50V R227 X202:6 SM6T15CA A3 Size 100nF U214:B 8 LM274 LM274 U214:C P201 20k + - + R280 820K 4 C262 1nF 12 13 5 6 R222 1k 9 10 D220 BAS85_2 C219 100nF U206:C Vcc GND C215 1nF 4 5 3 2 6 C246 4u7F 25V U209:A 1 R275 +5V 8 Drawn by 100nF C214 C265 R220 + LM358Y 10K 5,6K R276 DZ201 BZX84C6V8 11 Monteras ej!!! R221 3 2 - R279 4,7K C218 100nF U214:E Vcc GND 10k 100n C283 100n +10V 4 R309 2k2 C284 4 C263 Filename 100nF R310 24.9k 24.9k C208 100nF U209:C Vcc GND LM393 + VR202 KTY 81-220 VR201 KTY 81-220 +12V 8 LM274 3,3k AWM 2100V SENSOR C231 VR201 Monteras ej! 1uF/35V C282 100n LM358Y Monteras ej!!! R281 BREAS MEDICAL 7–18 Circuit diagram – MDA board 2, rev 3 7.6.9 PV 101+/PV 102+ Service Manual ELECTRONICS C266 C267 +35V 8 +12V X202:4 2,7k R257 LM358Y 1uF/35V 4 LM393 GND Vcc 8 +V C226 U200:C Vcc GND LM393 100nF C252 0,1uF +35V BAS32_2 4 4,7K R243 4,7K R244 4,7K R242 C236 330uF/25V + +5V D221 100nF U207:C 2,2K 10K Mont. ej! VR200 C247 3k3 R274 3 R260 8 2 DR212 VR200 Monteras ej 1 C224 R259 Vcc GND U212:C 100nF C276 4 10nF C243 4,7K 3,3K R241 1,82K C225 R268 3 1 DR211 2 R269 10nF X205:2 X205:3 C223 C275 100nF X205:1 X205:8 + NE555 1k 6 5 2 3 - 4k7 R313 U207:A LM393 LM393 5 6 1 7 470K R318 + - BAS32_2 D238 MC33035PY BRAKE ENABLECS DIR CSIN 60/120 EAIN EAOUT GND CSLIN OSC SPEED AT BT CT BAS32_2 FAULT AB BB CB U205 Vc Vcc SA SB SC VREF U207:B +5V - + + 18k R263 23 7 3 22 12 13 16 10 11 18 17 4 5 6 8 +12V U212:B 7 LM358Y 15 9 14 21 20 19 2 1 24 +35V + 4,7uF/16V C232 4,7uF/16V C233 C230 0,33uF 10nF C228 R316 R317 10nF C200 D210 10R D235 3,3nF 47K 47K 3,3K R285 10K R247 +V 100R R239 100R R288 + + - + 47k R295 BAS32_2 47k R297 220k R296 6 5 Q213 BC847B C242 R301 LM393 7 U200:B +V D237 BAS32_2 Q205 BZX84C10 D218 Q206 R300 R299 +35V RFD16N06LESM RDF16N06LESM 10K ? ? R264 BAS32_2 D239 R246 330pF TRIG Vcc R THR DIS OUT GND CON R261 R314 R315 0,47u 12K 15K 1K 50k P205 +12V 3,3K R286 BAS32_2 D232 2 3 100R R289 - + Q212 BC847B 10K R278 D233 C241 1k5 R272 LM393 1 U200:A 82K R277 BAS32_2 0,22R 10K C222 BAS32_2 D209 + C272 R240 100nF R249 R251 R262 C234 D217 RDF16N06LESM Q207 Q208 RFD16N06LESM 3 + - 22R G Q217 MTD20N06HDL R294 + S D +35V U212:A 1 LM358Y 4,7uF/16V 2 100K BZX84C10 R298 +12V 330pF 8 6 3 5 R266 R267 10K 4k7 U208 4R7 Mont. lodsida 2 4 7 1 100k 8k2 D223 R302 D224 R271 D211 BZX84C10 10K R258 BZX84C10 2k2 BAS32_2 R273 1K R250 10K 3k3 R254 BAS32_2 D234 R253 4R7 Mont. lodsida 100R R290 3,3K R287 R308 R307 R306 Q211 BC847B 3*4k7 +5V DO-214AB(SMC) D227 C251 C278 Q209 BZX84C10 D216 Filename Drawn by A3 Size Title 1 2 3 4 1 2 3 4 U211 8 7 6 5 1k R303 C248 +35V C279 100nF 100nF Axel O cad057a_R3.sch Number MDA Sheet Date Rev X202:13 X204:2 X204:1 X201:3 X201:5 X201:4 X202:14 X201:2 X201:8 X201:6 X201:7 X201:1 100nF 1 of 2003-01-14 PV101/102 Upgrade MDA 100nF +5V RFD16N06LESM Q210 +35V 8 7 6 5 BAS32_2 D236 MICRO_PIC12C67 100nF RDF16N06LESM 10K +5V 1nF C269 R304 +5V D222 1nF C249 +5V BZX84C10 1nF 1nF C270 10k 2k R305 +12V D228 D229 R252 DO-214AB(SMC) D230 C240 DO-214AB(SMC) D226 DO-214AB(SMC) D231 330pF DO-214AB(SMC) BREAS MEDICAL DO-214AB(SMC) 1nF C250 C268 C277 1nF C274 Doc. No. 003259 En Issue: X-1 C273 10uF 35V 3 3 PV 101+/PV 102+ Service Manual ELECTRONICS 7.6.10 Circuit diagram – MDA board 3, rev 3 7–19 10K R265 ELECTRONICS 7.7 7–20 PV 101+/PV 102+ Service Manual Component location drawings – Circuit boards BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7.7.1 ELECTRONICS Component location drawing – CPU board For component lists, refer to section 7.8.1. U102 C125 C116 C117 C123 C118 C107 C119 C120 C121 C122 C126 C127 C124 1 + C131 R143 R150 R142 R149 R141 R148 R140 R147 R139 R146 R138 R145 R137 R144 R136 R135 XTAL100 R101 U100 U119 C115 C133 C110 C134 R103 R104 R105 R106 C139 R100 R107 C101 C100 X100 C132 C105 C104 C102 C138 U105 U104 14 1 U101 1 C103 U106 C128 LED104 LED105 R121 Q101 R120 R125 Q102 R124 R122 LD106 C135 R117 R127 R126 R175 R174 R173 R172 R171 R170 R169 R168 R159 R158 R157 R156 R155 R154 R153 R152 C129 LED106 Q104 R123 R182 R181 R180 R179 R178 R177 R176 R167 R166 R165 R164 R163 R162 R161 R160 R151 U116 Q103 C109 LD105 C140 R118 LD107 + R115 U114 R102 OK104 U118 U115 C113 C136 C137 + + U113 U112 OK101 C112 R132 R133 LED100 R113 R114 LD103 R116 R128 LD102 LED101 C114 LD104 LED102 LED103 U117 R129 C130 R130 C108 R111 R110 Q100 R119 R112 LD100 R109 R108 R134 OK103 OK102 LD101 RN101 RN102 RN103 RN104 U109 RN105 RN106 RN107 + RN100 R131 C111 U110 U108 U107 2960 U111 OK100 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–21 ELECTRONICS 7.7.2 PV 101+/PV 102+ Service Manual Component location drawing – Power board, upper side For component lists, refer to section 7.8.2. CN2 R90 L3 Q7 D9 D22 D20 L2 + + + C24 C51 C27 1 R36 C4 TR3 R43 R82 R83 U2 C23 C18 C46 TR2 R32 C15 + C19 R86 Q12 Q11 C3 C45 0 AL + R28 R85 2959 4 D1 + R40 R37 Q5 R103 C33 + 1 C21 C6 C60 L4 TR1 S3 C13 L1 CN3 R8 C32 + C1 C12 ~ BR1 - TR4 R14 + ~ C6 BREAS MEDICAL CN1 R6 C10 C8 C7 7–22 Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7.7.3 ELECTRONICS Component location drawing – Power board, under side For component lists, refer to section 7.8.2. R15 D3 Q6 R58 R50 TP3 R1 TP5 U1 R51 C5 R52 R57 C55 R61 R5 R9 R16 R11 R12 + C56 R49 C53 R48 C59 C54 R59 U4 R2 R89 C64 R7 R88 R3 C2 R17 U9 C58 C52 R4 R60 + R10 D1 D12 R13 R55 R54 C28 D2 D18 C63 R87 C47 C57 C50 C29 R62 C26 C22 C31 R84 D19 R63 + R79 C37 R67 R78 R75 Q14 R93 R94 1 C44 R64 R95 Q10 R20 Q9 R74 Q8 R96 D11 C43 R70 + + R21 D15 C11 Q15 R8 1 Q1 D5 D7 D10 C42 + R73 R72 Q4 C38 R69 C41 C39 C40 R65 D21 D13 K1 R101 R102 C16 C17 + C66 Q17 R104 R105 R100 R99 R97 Q18 Q13 R56 R98 Q16 R53 C35 D8 R30 R18 R19 R26 C14 C36 R66 R34 U3 R27 C9 R39 D6 C20 R31 R29 C34 U7 Q2 R35 R33 R71 R68 Q3 R25 R22 R24 D4 D16 R38 D17 R23 R91 R76 C65 C25 R42 R77 R80 R92 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–23 ELECTRONICS 7.7.4 PV 101+/PV 102+ Service Manual Component location drawing – MDA board For component lists, refer to section 7.8.3. C230 C249 U208 C228 R264 + C232 R242 C251 R244 D210 D209 R262 C250 R243 C269 D222 R253 R287 Q211 X201 Q209 Q210 D216 C268 C240 R252 Q207 R286 Q212 R250 Q208 D217 C270 D223 C248 C241 U210 + + C265 DR207 VR201 R275 R276 U209 R282 R283 X206 C221 R279 D220 R280 01 C220 D204 R281 C2 R234 R212 U203 + R213 Q201 D203 D219 DZ201 R233 R292 200 DR201 R256 C210 R293 DR210 DR202 Q215 + R291 TIMER200 C253 C236 D215 DR205 DR204 DR206 D213 Q216 D212 X200 C262 + P205 D225 Q203 REL200 R236 DR208 DR203 X204 DR209 R257 R235 D208 D207 C238 C208 P203 G P204 G P201 C212 R206 R214 C209 U201 R216 R217 Q214 C205 R203 + C207 R204 C211 R284 R232 R205 X207 C280 C202 O D206 C245 C272 Q202 C257 D214 D205 R314 R298 R277 C231 D233 C234 R278 D221 R313 D238 R274 C275 0 R259 S20 C259 PZ200 C254 R300 R299 C244 R241 C224 R308 R307 R306 R263 R260 U212 R301 R239 C222 R261 R219 R218 C281 C278 U207 C226 D234 U205 U213 D235 U211 C277 D236 R316 C276 R240 R317 R302 R288 R289 C279 R304 R305 R290 C242 R285 R303 C200 VR200 C243 D239 R246 R268 + C233 C225 D224 Q213 R269 R318 D218 C247 R315 Q205 Q206 R247 R249 C204 C203 C206 R207 R310 R309 R215 C255 C213 C237 1 C261 C258 R229 R230 U206 R228 C219 R231 C264 P202 O R202 X203 C263 U202 R270 R210 R211 R209 R208 X205 C223 C266 R227 C271 U214 X202 C217 C216 R221 R224 C215 + C214 C274 D230 R222 R223 R225 D229 D228 R220 C260 C282 C218 R200 Q200 C256 C284 VR202 C267 R312 D200 R238 R226 D202 R201 R255 R237 DR211 C283 R311 DR212 C246 14 D227 D231 D226 D237 R266 R254 R258 R251 SENSOR201 C252 U200 R295 D232 R297 R267 SENSOR200 R272 D211 R271 R296 R294 7–24 R265 Q217 C273 R273 + BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual ELECTRONICS 7.8 Component lists – Circuit boards 7.8.1 Component list – CPU board Reference Description Quantity C100-103, 107-109 CAPACITOR 100 nF X7R 27 C104, 105 CAPACITOR 22 pF NPO 2 C110-113, 140 CAPACITOR 10 uF/6V 5 C116-118, 120, 121 CAPACITOR 1 nF NPO 6 LD100 LED Red 3mm 1 LD101, 106 LED Green 3mm 2 LD102, 104, 105, 107 LED Green L113GDT 2x5mm 4 LD103 LED Red/Green Circular 1 LED100-103 TDSG1160 7-seg. 4 LED104, 105, 106 Bargraph PV 101+/PV 102+ 3 OK100-104 SWITCH 3ESH9-10,4 5 Q100-104 TRANSISTOR BC847B 5 R100, 101 RESISTOR 3k3 1% 2 R102 RESISTOR 10k 1% 1 R103-106 RESISTOR 4K7 1% 4 R107, 109, 119, 133 RESISTOR 1k 1% 4 R108, 110, 113-118 RESISTOR 470R 1% 12 R111 RESISTOR 100k 1% 1 R112 RESISTOR 4k7 1% 1 R120-123 RESISTOR 47R 1% 4 R124-127 RESISTOR 2k2 1% 4 R128 RESISTOR 560R 1% 1 R134 RESISTOR 470K 1% 1 R135-143, 151-159 RESISTOR 20k 1% 27 R144-150, 160-166 RESISTOR 10K 1% 21 RN100-107 RN RESNET MNR34-J5AJ-471 8 U100 IC 24LC01 WSOITR 1 U101 IC DS14C232CWM SO16W (232CB) 1 U102 IC HD63C03YRF 1 (U104) IC PLCC32 SMD 1 U105 IC GAL16V8H-25SC/4 1 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–25 ELECTRONICS 7.8.2 7–26 PV 101+/PV 102+ Service Manual Reference Description Quantity U106 IC 74HC00 1 U107, 108, 110, 111 IC 4511 4 U109, 115, 116 IC 74HC574 3 U112 IC 4093 1 U113 IC 4027 1 U114, 117, 118 IC 74HC273 3 U119 IC MAX699CWE 1 XTAL100 XTAL 10 MHz HC49/4H 1 Component list – Power board Reference Description Quantity BR1 BRIDGE DB154G 1 C1 CAPACITOR 220nF PME271 1 C2, 20, 26, 28, 29, 34 CAPACITOR 100nF X7R 17 C3, 4 CAPACITOR 2,2nF Y1 DE1210E222M 2 C5 CAPACITOR 10u/16V 1 C6 CAPACITOR 68nF PME271 1 C7, 8, 10, 12, 15 CAPACITOR 1nF/400V 5 C11, 37 CAPACITOR 47u/16V 2 C13 CAPACITOR 100uF 400V 1 C14, 36, 41 CAPACITOR 10nF X7R 3 C16, 25, 39 CAPACITOR 2,2nF X7R 3 C17 CAPACITOR 330pF NPO 1 C18 CAPACITOR 33u/50V 1 C19 CAPACITOR 100nF/400V 1 C21 CAPACITOR 10uF 50V 1MOD RSH 1 C22 CAPACITOR 330p/500V 1 C23 CAPACITOR 220pF 400V 1 C24, 27, 32, 33 CAPACITOR 1000uF 50V 4 C31, 50, 63, 64 CAPACITOR 220pF 500V 4 C40, 42 CAPACITOR 680p NPO 2 C43 CAPACITOR 10u/20V 1 C45, 46, 60, 61 CAPACITOR 4,7nF 500V XY 4 C51 CAPACITOR 100u/50V 1 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual ELECTRONICS Reference Description Quantity C58 CAPACITOR 0,33u/35V 1 C65 CAPACITOR 22u/10V 1 C66 CAPACITOR 4,7uF/16V 1 CN1 CONNECTOR 3P 1 CN2 CONNECTOR 8P 1 CN3 CONNECTOR 2P 1 D1 DIODE BZX84C3V6 1 D2, 3, 18 DIODE BZX84C5V1 3 D4 DIODE BZX84C24 1 D5 DIODE RS3K SMC 1 D6, 8, 16, 17 DIODE BAS21 4 D7, 10, 11, 19 DIODE BYD37M 4 D9, 20, 22 DIODE BYT12PI-1000 3 D12 DIODE BZX84C10 1 D13 DIODE BZX84C22 1 D14 DIODE BYW29F200 1 D15 DIODE BZX84C6V8 1 D21 DIODE BZX84C20 1 L1 INDUCTOR 220uH HM50-221K 1 L2 INDUCTOR 22uH HM50-220K 1 L3 INDUCTOR 330uH KM5 1 L4 INDUCTOR TSL1112RA 1 Q1, 3, 6, 8, 9, 14, 17, 18 TRANSISTOR BC817-40 5 Q2, 4, 10, 13, 16 TRANSISTOR BC807-40 5 Q5 TRANSISTOR SPP20N60S5 1 Q7 TRANSISTOR IRF5305 TO220 1 Q11, 12 TRANSISTOR IRFIZ44N TO-220 2 Q15 TRANSISTOR MTD20P06HDL 1 R1, 4, 5, 9, 10, 13, 20, RESISTOR 4k7 1% 11 R2, 18 RESISTOR 3k3 1% 2 R3, 27, 31 RESISTOR 8K2 1% 3 R11, 12, 29 RESISTOR 1K5 1% 3 R14 VARISTOR S14K300 1 R16 RESISTOR 15K 1% 1 R17 RESISTOR 470R 1% 1 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–27 ELECTRONICS 7–28 PV 101+/PV 102+ Service Manual Reference Description Quantity R19 RESISTOR 18K 1% 1 R21, 67 RESISTOR 56R 1% 2 R22, 34, 72, 73, RESISTOR 1k 1% 6 R23 RESISTOR 3k9 1% 1 R25 RESISTOR 270k 1% 1 R26, 60, 71, 92, 98 RESISTOR 47k 1% 5 R28, 103 RESISTOR 220k 0,6W 2 R30 RESISTOR 68R 1% 1 R32 RESISTOR 5K6 2W 1 R33 RESISTOR 150R 1% 1 R35, 54 RESISTOR 22k 1% 2 R36 RESISTOR 270R 0,6W 1 R37, 40 RESISTOR 0R68 1W PR01 2 R38, 76, 77 RESISTOR 10R 1% 3 R39 RESISTOR 33R 1% 1 R42, 52, 62, 75, 78, 79 RESISTOR 100k 1% 15 R43 RESISTOR 4K7 2W 1 R48, 57, 64, 65, 87 RESISTOR 10k 1% 7 R50, 51, 70 RESISTOR 6K8 1% 3 R55, 59, 84 RESISTOR 2k2 1% 3 R58 RESISTOR 10k NTC B57621C103 1 R6 RESISTOR B57236-S200-M NTC 1 R61, 94 RESISTOR 470K 1% 2 R63, 91 RESISTOR 47R 1% 2 R66, 105 RESISTOR 68k 1% 2 R69 RESISTOR 12K 1% 1 R7, 15 RESISTOR 680R 1% 2 R8 RESISTOR 1M 0,6W 1 R82, 83 RESISTOR 0R1 0,6W 2 R85, 86 RESISTOR 47R O,6W 2 R88, 89 RESISTOR 100R 1% 2 R90 RESISTOR RXE185 / 3.5 A fuse (wire type) 1 S3 FUSE HOLDER Schurter 5x20m 1 TR1 TRANSFORMER 12uH PLA10 1 TR2 TRANSFORMER ETD34-3F3 1 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 7.8.3 ELECTRONICS Reference Description Quantity TR3 TRANSFORMER ETD29 1 TR4 TRANSFORMER 7uH PLA10A7720R7D02 1 U1 IC LM324Y 1 U2, 6, 8 IC CNY17F-3 DIL6 3 U3 IC UC 3845D 1 U4 IC LM393D 1 U7 IC UC3825DW 1 U9 IC LM78L05ACM 1 Component list – MDA board Reference Description Quantity C200 CAPACITOR 0,47u/25V 1 C201, 202 CAPACITOR 2,2nF Y1 DE1210E222M 2 C209, 232-234 CAPACITOR 4,7uF/16V 4 C210 CAPACITOR 10uF/6V 1 C213-217, 249-251 CAPACITOR 1nF NPO 11 C220 CAPACITOR 220u/50V 1 C221 CAPACITOR 0,47F FSOH 1 C222, 224, 225 CAPACITOR 10nF X7R 3 C223 CAPACITOR 10nF X7R 1 C226 CAPACITOR 0,1uF/100V MKT 1 C228 CAPACITOR 3,3nF X7R 1 C230 CAPACITOR 1uF/50V 1 C231, 243 CAPACITOR 1uF/35V 2 C236 CAPACITOR 330uF/25V CV-GX 1 C237 CAPACITOR 33uF/16V 1 C238, 245 CAPACITOR 0,1uF/50V X7R 2 C240-242 CAPACITOR 330pF NPO 3 C244 CAPACITOR 100uF/25V CV-GX 1 C246 CAPACITOR 4,7uF/35V ELYT D55 1 C273 CAPACITOR 10uF/50V MVKF55 1 C282-284, 203, 204 CAPACITOR 100nF X7R 35 D200, 202, 204, 209 DIODE BAS32 15 D203, 212, 213, 219 DIODE BAS85 5 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–29 ELECTRONICS 7–30 PV 101+/PV 102+ Service Manual Reference Description Quantity D205-208, 214 DIODE SM6T15CA 5 D215, 225 DIODE BYS10-45 2 D216-218, 222-224 DIODE BZX84C10 6 D226-231 DIODE RS3K SMC 6 DR201-205, 208 INDUCTOR DSS306-91FZ103N100 9 DR206, 209 INDUCTOR 330uH HM50-331K 2 DR207 INDUCTOR BL01 RN1-A62 1 DZ201 ZENER DIODE BZX84C6V8 1 P201, 204 POTENTIOMETER 20K TYP3224J 2 P202 POTENTIOMETER 1K TYP3224J 1 P203 POTENTIOMETER 5K TYP3224J 1 P205 POTENTIOMETER 50K TYP3224J 1 PZ200 BUZZER OBO-1206C 1 Q200 TRANSISTOR BC857B 1 Q201, 202, 211-213 TRANSISTOR BC847B 7 Q203 TRANSISTOR LM2575S-5 1 Q205-210, 214 TRANSISTOR RFD16N06LESM 7 Q217 TRANSISTOR MTD20N06HDL 1 R200, 233, 234 RESISTOR 3k3 1% 3 R201, 256 RESISTOR 1k 1% 2 R202, 214, 288-290 RESISTOR 100R 1% 5 R203, 216, 225, 226 RESISTOR 270R 1% 4 R204 RESISTOR 39K 1% 1 R205, 222, 227, 229 RESISTOR 1k 1% 11 R206, 215, 235-238 RESISTOR 10k 1% 15 R210 RESISTOR 1k6 1% 1 R211, 212 RESISTOR 220R 1% 2 R213 RESISTOR 680R 1% 1 R218 RESISTOR 560R 1% 1 R220, 221 RESISTOR 24K9 2 R228, 231, 260, 271, 310 RESISTOR 2k2 1% 5 R239 RESISTOR 100R 1% 1 R240 RESISTOR 0,22R 3W AC03 1 R241-244 RESISTOR 4k7 1% 4 R246, 247, 249, 250 RESISTOR 10k 1% 6 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual ELECTRONICS Reference Description Quantity R255, 268, 281, 285 RESISTOR 3k3 1% 12 R257 RESISTOR 2k7 1/2W 5% 1 R263, 209 RESISTOR 18K 1% 2 R264, 270, 292 RESISTOR 10R 1% 3 R266, 283, 298 RESISTOR 100k 1% 3 R267 RESISTOR 8K2 1% 1 R269 RESISTOR 1k82 1% 1 R276 RESISTOR 5k6 1% 1 R277 RESISTOR 82k 1% 1 R279, 219, 251, 313 RESISTOR 4k7 1% 8 R280 RESISTOR 820k 1% 1 R293, 272 RESISTOR 1K5 1% 2 R294 RESISTOR 22R 1% 1 R295, 297, 316, 317 RESISTOR 47k 1% 4 R296 RESISTOR 220k 1% 1 R305 RESISTOR 2k 1% 1 R312 RESISTOR 5k1 1% 1 R314 RESISTOR 12K 1% 1 R315 RESISTOR 15K 1% 1 R318 RESISTOR 470K 1% 1 REL200 RELAY RT114024 1 S200 POLYFUSE RXE020 1 U200, 206, 207 IC LM393D 3 U201, 214 IC TLC274CDR 2 U202, 203 IC LM317LM 2 U205 IC MC33035DW 1 U208 IC NE555 1 U209, 212 IC LM358D 2 U210 IC LM2597M-12 1 U213 IC LM317 1 VR202 SENSOR KTY81-220 1 X201 CONNECTOR ANGLED 8P 1 X202 CONNECTOR 1x14POLE 1 X203, 204 CONNECTOR 2P 2 X205 CONNECTOR 8P 1 Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 7–31 ELECTRONICS 7–32 PV 101+/PV 102+ Service Manual Reference Description Quantity X207 CONNECTOR ANGLED 2P 1 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 8 FAULT TRACING FAULT TRACING This chapter contains a fault-tracing table that may be used when trouble-shooting the device. 8.1 Fault-tracing table If the PV 101+/PV 102+ does not work properly try to identify the problem in the table below. Check the possible causes and carry out the suggested remedial actions. Symptom Possible cause Remedial action See ref. The device alarms at start-up. • The power cord is not properly connected. • Connect the power cord. 2.4.2 • Mains fuse has blown. • Replace the fuse. 7.1 • Leakage from patient circuit/ mask. • Check the tubes and mask for leakage. 2.6.6 • Internal leakage. • Check the tubes inside the machine. 2.5.3 • Dirty filters. • Replace or wash the filters. 2.6.5 The ventilator does not give adequate pressure/volume. 1 Adjust voltage to transformer. 2 Replace MDA board or pressure sensor. 4.3 • Trigger settings not correct. • Adjust settings. See PV 101+/ PV 102+ Operating manual • Tube to flow sensor has worked loose. • Check tube and connectors. 7.4.3 • Pressure sensor faulty. The PV 101+/ PV 102+ does not trigger. • Flow sensor faulty. • MDA board faulty. Alarms for mask-off (PV 102+ only). • Leakage from patient circuit/ mask. Doc. No. 003259 En Issue: X-1 1 Adjust voltage to pressure sensor. 2 Replace MDA board or flow sensor. • Check the tubes and mask for leakage. BREAS MEDICAL 3.7 4.5 3.7 2.6.6 8–1 FAULT TRACING 8–2 PV 101+PV 102+ Service Manual BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 9 APPENDICES APPENDICES This chapter contains the engineering change history of the PV 101+ and the PV 102+. Included in the chapter is also a service record and a service report. These pages can be copied and used for completing the reports. 9.1 Engineering change history – PV 101+ The table below lists the engineering changes made to the PV 101+ by serial number. From serial no. Changes 9.2 R060334 Software rev 3. Hardware MDA rev. 2, PSU rev. 3, CPU rev. 3 Production start. R090177 Software upgrade rev 4. Ref. internal improvement no. 337: The PV 101+ incorrectly turned itself off 15 seconds after ramp activation when there was no patient circuit leakage. R250001 (mandatory field upgrade from R020296 to R230389) Replacement of the R90 component on the power board (ref. TBD33). A 3.5 A fuse (wire link type) has replaced the former PTC fuse, in order to avoid risk of disturbances due to ambient temperature. Engineering change history – PV 102+ The table below lists the engineering changes made to the PV 102+ by serial number. From serial no. Changes R060335 Software rev 3. Hardware MDA rev. 2, PSU rev. 3, CPU rev. 3 Production start. R090279 Software upgrade rev 4. Ref. internal improvement no. 337: The PV 102+ incorrectly turned itself off 15 seconds after ramp activation when there was no patient circuit leakage. R250001 (mandatory field upgrade from R020296 to R230389) Replacement of the R90 component on the power board (ref. TBD33). A 3.5 A fuse (wire link type) has replaced the former PTC fuse, in order to avoid risk of disturbances due to ambient temperature. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 9–1 APPENDICES 9–2 PV 101+/PV 102+ Service Manual BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual APPENDICES Service Record for BREAS PV 101+/PV 102+ Use a photocopy of this service record for the maintenance inspection described in chapter 2 in this service manual. Use the reverse side for comments and notes. Service Record No:__________ Registration Model:................................... Serial No................................... Inventory No. .................. Accessories: ............................................................................................................................... Delivery date:...................................... No. of operating hours: .............................................. Service started:................................... Signature: .................................................................. Service completed: ............................. Signature: .................................................................. Delivered, date: .................................. Signature: .................................................................. See instruction ref. Check OK Preparing inspection Verifying the components and the installed software Initial recording Markings Information from the patient/user Validity of the documentation 2.3.1 2.3.2 2.3.3 2.3.4 2.3.5 ......... ......... ......... ......... ......... 2.4.1 2.4.2 2.4.3 ......... ......... ......... Cleaning the inside of the machine Inspecting the internal wiring Securing the components Checking the power supply Checking Electrical safety measurements 2.5.2 2.5.3 2.5.4 2.5.5 7.5 ......... ......... ......... ......... ......... Blower Assembly (15,000 hours) Checking if blower unit should be replaced 3.5 ......... Calibration (12 months/15,000 hours) Calibrating the pressure sensor Calibrating the flow sensor Calibrating the high pressure limit 4.3 4.5 4.6 ......... ......... ......... External inspection Visual inspection for external damage and wear Checking the power supply connection Short function check Internal inspection Continued on next page Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 9–3 PV 101+/PV 102+ See instruction ref. Check OK Final inspection before delivery Function check Checking the low pressure leakage alarm (PV 102+) Checking the trigger function Checking the pressure and rate Replacing the filters Checking the accessories Adjusting the settings for the patient 2.6.1 2.6.2 2.6.3 2.6.4 2.6.5 2.6.6 2.6.7 ......... ......... ......... ......... ......... ......... ......... Notes ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual 9.3 APPENDICES Returning products to BREAS You may need to return the device or any components or accessories to BREAS, for example, for service, warranty, upgrade or repair. If so, follow the instructions below to ensure that the correct action is taken and to avoid unnecessary delays. 1 Pack the product in its original packaging. If this is not available pack the product in packaging suitable for transporting the product to BREAS. 2 Take a photocopy of the delivery report on the next page. 3 Fill out the Customer information part of the service report and pack it together with the product to be returned. The service report will be completed by BREAS and returned with the product. Product damage caused by poor packaging or during transport is not covered by the factory warranty. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 9–5 APPENDICES 9–6 PV 101+/PV 102+ Service Manual BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Report for BREAS PV 101+/PV 102+ BREAS Ref. No:__________ Customer information Customer name:................................................................................................................... Customer address:............................................................................................................... ............................................................................................................................................. Phone:.................................................................................................................................. Reference person:................................................................................................................ Customer reference number: ............................................................................................... Product information Model: ....................... Serial No:.......................... Operating hours: ........................h Error /Complaint / Accessories ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. Doc. No. 003259 En Issue: X-1 BREAS MEDICAL PV 101+/PV 102+ Information from BREAS Date received by BREAS:...................... Repair Warranty Signature ................................................... Update Charge Other Action taken: _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ Parts used: Pcs: Price: _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ Operating hours from BREAS:_____________h Date returned to customer: ________________ Signature:____________________ Notes: _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ _______________________________________________________________________ BREAS MEDICAL Doc. No. 003259 En Issue: X-1 PV 101+/PV 102+ Service Manual APPENDICES Notes ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 9–9 APPENDICES PV 101+/PV 102+ Service Manual Notes ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ ___________________________________________________________________ 9–10 BREAS MEDICAL Doc. No. 003259 En Issue: X-1 10 PV 101+ OPERATING MANUAL BREAS MEDICAL BREAS MEDICAL 11 PV 102+ OPERATING MANUAL BREAS MEDICAL BREAS MEDICAL