SERVICE MANUAL MV2000 MEKICS VENTILATOR MEKICS Co., Ltd Industrial Zone, 861-10 Taegye-dong, Chunchon-si, Kangwon-do, Korea TEL + 82 33 264 5926 www.mek-ics.com FAX +82 33 264 5999 E-mail : sales@mek-ics.com MV2000 SERVICE MANUAL VER 1.1 November, 2008 - TABLE OF CONTENTS 1. General information …………………………………………. 1 2. Configuration of MV-2000 …………......…………………… 6 3. Structure of MV-2000 ………..………………….…….. 12 (1) Front picture and the explanation …………………….. 12 (2) Back side picture and the explanation ………………… 14 (3) Left side picture and the explanation …………………… 16 (4) Right side picture and the explanation (5) Keypad ………………… 17 ……………………………..……………………. 18 4. Power ON …………………….…………………………… 22 5. Specifications ………….…………………………………… 25 6. Description of each board ………………………………… 27 (1) System Block Diagram (over all system) (2) DSC Board ……..……………. 27 …………………………….….………………… 28 (3) PNEUMATIC Board …………………………………………. (4) DRIVER Board ……………………………………..…… 41 (5) SENSOR Board …………………………………………..….... (6) KEY Board 35 43 …………………………………………..….... 45 (7) POWER Board(AC–DC) …………………………………….. 46 7. Assembly & Disassembly ……………………………… 49 (1) Product Disassembly ……………………………………….. 49 DRIVER BOARD Separation …………………………. 59 SENSOR BOARD Separation …………………………... 60 PNEUMATIC BOARD Separation SOLENOID VALVE Separation II ..……………………... 61 ………………………... 62 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (2) Monitor’s Separation …………………………………………. 64 Monitor’s Separation ………………......………………… 64 Monitor Main Frame Separation INVERTER BOARD Separation …………………………….. DSC BOARD Separation TFT-LCD Separation KEY BOARD Separation ALARM and NORMAL LED Separation ………..………………. 66 69 ……………………………… 70 ………………………………… 71 ……………………………… 72 ………….…. 73 8. Voltage Control method ……………………………….… 74 (1) Each Board's Voltage control method ………………………… 74 1) Digital Board’s Voltage Control ..……………………………… 74 2) Sensor Board’s Voltage Control ..…………………..……………. 3) O2 Sensor's voltage control 75 .………………………………… 75 (2) Product inspection method .………………………………………. 76 1) Sol Valve on/off TEST .…………………………………..…… 76 2) PV1 Adjustment ……………………………………..….... 77 3) LS1 Adjustment ……………………………………..….... 78 9. TROUBLE SHOOTING …………………………………….. 79 (1) LCD Screen Badness …………………………………… 80 (2) Solenoid Valve’s Operation Badness …………………….. 82 (3) Unable to measure VENT MODE …………………………. (4) KEY Operation Badness 83 ………………………………… 84 (5) Power Deficiency & Unable to turn on the device ………. 85 10. COMMUNICATION ……………………………………... 87 11. Part List ………………………………………..………….... 88 III MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 1. General information CE Marking Information Standard List >>> 1. Electrical Safety EN60601-1 : +A1 : 1993, +A2 : 1995, +A13 : 1996 Medical electric equipment Part1: General requirements for safety 2. EN60601-1-2 : 2001 (IEC 60601-1-2:2001) Electromagnetic compatibility-Requirements and tests 3. EN60601-2-12:2001 Medical Electrical Equipment: Part 2-12: Particular requirements for the safety of lung ventilator Critical care ventilator 4. EN61000-3-2:2000(IEC61000-3-2:2000) Electromagnetic compatibility(EMC)-Pt.3 : Limits for harmonic-current emissions (Equipment input current up to and including 16A) 5. EN61000-3-3:1995+A1:2001(IEC61000-3-3:1994+A1:2001) Electromagnetic compatibility(EMC)-Pt.3 : Limits-S.3 : Limits for voltage fluctuations and flicker in low Voltage supply systems for equipment with rated current up to and including 16A. 1 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 Warning, Caution Symbol Warning Warns of conditions and situations which, if not taken care of or avoided, could cause damages to people or to the equipment Caution Warns of conditions and situations which, if not taken care of or avoided, could cause damages to people or to the equipment Note Informs the reader about the relevant facts and conditions: draws the attention to important information however this does not necessarily cause damages to people or to the equipment. 2 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 General Precaution on Environment Avoid placing in an area exposed to moisture. Do not touch the equipment with wet- hand. Avoid exposure to direct sunlight Avoid placing in an area where there is a high variation of temperature. Operating temperature ranges from 10°C to 40°C. Operating humidity ranges from 30% to 75%. Avoid in the vicinity of Electric heater Avoid placing in an area where there is an excessive humidity rise or ventilation problem. Avoid placing in an area where there is an excessive shock or vibration. Avoid placing in an area where chemicals are stored or where there is in danger of gas leakage. Avoid dust and especially metal material into the equipment Do not disjoint or disassemble the equipment. MEKICS Co., Ltd. takes no responsibility of it Power off when the equipment is not fully installed. Otherwise, equipment could be damaged. 3 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 General Precaution on Electric Safety Warning Before operate this equipment, please check out followings; - Be sure that power supply line is appropriate to use. (Power Input: 100 - 240V AC). - Be sure that the entire connection cable of the system is properly and firmly fixed. Note Note: The equipment should not be placed in the vicinity of electric generator, X-ray, broadcasting apparatus to eliminate the electric noise during operation. Otherwise, it may cause incorrect result. Self-power line is important for MV2000. To use the same power source with other electric instruments may cause incorrect result. Warning If MV2000 doesn’t operate normally or is damaged, please don’t use it to patient but contact your hospital engineer or your local representative. Equipment Connection Caution Doctors and nurses can be disposed the danger of uncontrolled electricity. This mentioned electricity can be made due to the electrical discrepancy between MV2000 and other electricity-conductable objects. In order to solve this problem, all configurations of other related equipment shall comply with the system standard EN 60601-1. 4 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 Maintenance and Cleaning MV2000 can be kept cleaning in many different ways. Use the following recommendations to avoid the damage or stain of the machine. 1) If the material (not approved material) that may cause damage to the product, the product will not be guaranteed even within warranty period. 2) Please, Check the main equipment and sensor carefully after its cleaning. If it is damaged or old, do not use it. 3) Clean the outside of equipment with clothes soaked warm water or alcohol once a month at least. 4) Do not use spray, thinner, ethylene or oxidizer. 5) Do not soak the machine or the probes into any liquid or detergent. Keep the machine or the probes away from any liquid. Also, don’t get wet any part of equipment nor accessories by water or any detergent. Caution Do not leave or throw away disposable sensor in the place caused damage. Be careful environment pollution. Caution Only the engineer from MEKICS or authorized maintenance service engineer can disassemble the product to chance Battery. If not, we can not grantee warranty service. . Caution Do not use MV2000 for 2 patients at the same time. 5 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 2. Configuration of MV-2000 (1) Components ① Ventilator (High pressure, Inspiration, Expiration) ② 10.4” TFT Color LCD Monitor ③ Cart ④ Circuit Arm : 1set 6 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ⑤ High Pressure Hose (O2, Air) ⑥ Inspiration Bacteria Filter (Disposable, Reusable) ⑦ Test Lung 7 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ⑧ Reusable Circuit (Silicone Hose(45cm, 60cm, 110cm)) ⑨ Tube connector(straight, Y-right angle) ⑩ Water Trap 8 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ⑪ Monitor VGA Cable ⑫ AC POWER Code ⑬ FUSES 2 Pcs (250V / 6.3A) ⑭ USER’S MANUAL 1 Copy 9 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL No VER 1.1 November, 2008 Name Explanation remark MEKANT as respiration assistant equipment for adult helps for normal breathing by ① setting breathing way, breathing volume, Ventilator breathing times, oxygen density, breathing rate etc in order to solve hypoxia to be appeared from emphysema of the lungs patients or respiratory disease patients. ② ③ ④ 10.4” TFT Color LCD Monitor Show the condition of various ventilators and patients. Cart Moving the system Support Arm Install Circuit Arm for patient. Oxygen High pressure hose ⑤ length:1.8m Connection hole to connect to main body using high pressure hose to approve O2 gas. length:3.5m Connection hole to connect to the main body, Air High pressure hose using high pressure hose to approve air gas. length:3.5m Device to filter the moisture in the Air. ⑥ ⑦ ⑧ filter is used at the place for exhaled air from Bacteria Filter patient’s body in order not to supply polluted air to patient. Test Lung Modeling of lung Respiration Tube(L:110Cm) Respiration Tube(L:60Cm) Between product and patient air inhalation and exhaust used tube Respiration Tube(L:40Cm) ⑨ Water Trap length:0.2m length:1.1m length:0.6m length:4.5Cm Connection product and tube, tube and tube Connector (L:6.5Cm) ⑩ ⑪ ⑫ Φ diameter: 72 length:0.4m Connector (L:4.5Cm) Connector Y (L:6.5Cm) Inside length:6.5Cm length:6.5Cm Remove water of air that is been hard on from patient. length:0.13m Monitor VGA Cable Power supply and Data transmission length:0.6m AC Power Cable power supply Cable length:1.9m 10 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL ⑬ ⑭ FUSES 2 Pcs (250V/6.3A) USER’S MANUAL VER 1.1 November, 2008 Protection of System operator’s manual explains the way of installation and usage for MEKANT Table 22-1. Component Description 11 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 3. STRUCTURE OF MV-2000 (1) Front picture and the explanation ⓑ LCD MONITOR ⓐ CIRCUIT ARM SOCKET ⓒ KEY FRONT ⓐ CIRCUIT ARM SOCKET ⓗ NEBULIZER PORT ⓓ INSPIRATION PORT ⓖ CARRIAGE CARRIAGE GRIP ⓕ EXPIRATION LEVER ⓔ EXPIRATION PORT Figure 3-1. Front View 12 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL No. VER 1.1 November, 2008 Part Description ⓐ Circuit Arm Install Circuit Arm for patient. ⓑ LCD Monitor Show the condition of various ventilators and patients. ⓒ Key Front Key button and encoder. ⓓ Inspiration Connector Joint hole to connect with patient. ⓔ Expiration Connector Joint hole to connect from patient. ⓕ Expiration Lever Lock or unlock lever for Expiration Assembly. ⓖ Carriage Grip ⓗ Nebulizer Port Connection to spray medicines Table 33-1. explanation Warning Be careful because heated exhalation assembly due to moisture during exhalation. 13 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (2) Back side picture and the explanation ⓐ SPEAKER ⓑ VGA ⓑ COMM. ⓑ TO PNEUMATIC ⓓ EXTERNAL BATTERY ⓔ POWER SWITCH ⓒ AC INLET ⓕ O2 INLET ⓚ COMM. ⓖ AIR INLET ⓙ TO MONITOR ⓗ TANK RELEAF ⓘ BATTERY DOOR Figure 3-2. Back side View 14 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL No. VER 1.1 November, 2008 Part Description ⓐ Speaker For alarm and sound. ⓑ VGA VGA cable connection hole to show patient condition on the other Monitor. COMM Cable connection hole to connect PC or central connection. To Pneumatic Connection hole to connect to the main body. ⓒ AC Inlet Connection hole for AC power input. ⓓ External Battery Connection hole for external battery. ⓔ Power Switch Power on/off switch for common use. ⓕ O2 Inlet Connection hole to connect to main body using high pressure hose to approve O2 gas. ⓖ Air Inlet Connection hole to connect to the main body, using high pressure hose to approve air gas. Device to filter the moisture in the Air. ⓗ Tank Relief Outlet to exhaust the over pressed air when the pressure of air tank is over than 40 PSI. ⓘ Battery Replacement Door to replace the internal battery. Door ⓙ To Monitor Connection for monitor. ⓚ COMM Connection hole for pneumatic communication. Table 33-2. explanation Warning You should be careful with battery’s polarity when it is exchanged. Products can be damaged when polarity is connected wrongly. It should be careful of battery’s terminal to be short on the metal part 15 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (3) Left side picture and the explanation ⓐ LOCKER Figure 3-3. Left side View No. ⓐ Description LOCKER : Locker to disassemble main body with cart. Pull to unlock and push to locker Table 33-3. explanation 16 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (4) Right side picture and the explanation ⓐ O2 CHANGE DOOR ⓑ INSPIRATION RELI RELIEF Figure 3-4. Right side View No. ⓐ ⓑ Description O2 CHANGE DOOR : Door to be used for exchanging O2 Sensor. INSPIRATION RELI RELIEF : Safety valve to exhaust by software, hardware and mechanically when the equipment is out of order. Table 33-4. explanation Warning Do not stop up safety valve to exhaust by software, hardware and mechanically when the equipment is out of order. 17 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (5) Keypad Figure 3-5. Keypad No Picture Description START SWITCH – If you press START switch for 0.5 sec, power would be on. ⓐ In case of STOP OFF you want , switch should be pressed for longer than 2 sec. MANUAL INSPIRATION : This function which is for a ⓑ MANUAL INSPIRATION ⓒ 100% O2 respiration according to Ventilator's default is operated once by pressing the key. 100% O2 : If you press this button, 100% O2 is provided for 2 minutes. NEBULIZER : It is acted once as setting value. This allows ⓓ NEBULIZER to nebulize during only inspiration and is stopped by pressing once again. 18 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 INS. PAUSE : Operating of Ventilator is stopped for a ⓔ INS. PAUSE ⓕ EXP. PAUSE ⓖ regular second after Inspiration Time. EXP. PAUSE : Operating of Ventilator is stopped for a VENT SETTING regular second after Expiration Time VENT SETTING : This changes Respiration mode & each parameter. But if RUN key or button is not pressed, modified value would be canceled.. ⓗ SETTING FREEZE : This is for stop of graph on the display. Stop ⓘ ⓙ ALARM SETTING : It is set alarm limit of each parameter. graph is measured after its FREEZE 2MIN ALARM 2 MIN : This makes ALARM remove for 2 minutes. ALARM RESET : This actives when alarm is reset. ⓚ RESET And previous alarm lamp is removed and the status of NORMAL will be. 19 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ACCEPT RUN : It is used when ⓛ Ventilator is working as its final value. If ACCEPT key is not pressed, setting value would not be accepted on the change of MODE and VENT SETTING. This is for the safe. ⓜ KEY LOCK : It is limit key of all operations. It can be set and removed in case of only pressing ACCEPT and CLEAR at the same time. Green LED is lighted ⓝ when it sets. CLEAR : It is used for clearance of menu when menu is controlled by Knob Encoder : Rotary switch for selecting ⓞ function on the pop-up menu. Table 33-5. explanation 20 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL A. No VER 1.1 November, 2008 LED Indicators Picture Description AC POWER ON LED : LED is on when machine is connected ⓟ to AC, otherwise, LED is OFF. BATTERY ON LED : Battery is charged as the machine is working when it is connected AC ⓠ — ┼ If it is not, battery ON LED is lighted because the machine is operating by internal battery. If both AC POWER ON LED and BATTERY ON LED are OFF, battery should be charged. EXT. ⓡ BATT. EXT. BATT : LED is on when the machine is connected to external battery, otherwise, LED is OFF. Table 33-6. explanation 21 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 4. POWER ON ⓐ INLET FUSE Figure 4-1. AC INLET Warning Prior to applying AC or DC to the ventilator, the line frequency has to be adjusted to the local standard otherwise it might causes damages to the apparatus. Power Cord has to be plugged in to the concentric plug on the ground. Replace covered wire in case of wear and tear. Please apply the ground cable in order to protect the patient and the apparatus. Maintain the same level of input & output voltage. No. Name Picture Description Pull the upper part of inlet toward you. Fuses are located where two upper arrows are and those are and those indicate ⓐ the lower 2 direction arrows when inserted. INLET FUSE Remove the existing fuse and insert the new fuse as shown in the picture after checking the rating. Close the inlet by doing the steps backward. Table 44-1. explanation 22 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL ▶ VER 1.1 November, 2008 Battery Operation It displays battery capability. This is displayed on the upper right hand corner of the LCD display. The battery capacity will be displayed in 5 main status. Battery Indicator Level Status Explanation Full Charging 2/3 Charging 1/2 Charging Low Charging In case of discharging status, the Discharging system will be automatically turn off to protect the battery 5 minutes later. Caution When Battery supply is low, there can be error in detecting vital-sign value. Please connect AC Adaptor when Low battery status is on. Warning Please contact our service center when you have problem with battery, do not disassemble it. 23 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ▶ Rotational control knob Rotating and Pressing the Encoder allows a user to navigate through and make changes to the display element. Left Rotating Right Rotating - Menu Down - Menu Up - Left Value - Right Value Click - Menu Screen - Menu confirmation & storing 24 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 5. Specifications Display & Key Panel Pneumatic Supply Display Panel with Key & Encoder Oxygen/Air Inlet Pressure 2.56~7.3Kgf/ Screen Size : 10.4” TFT Color LCD / 2.4~6.9 bar Resolution : 640 X 480 / 35~90 psi Special Key Functions ㎠ Pneumatic Specifications Mode - Manual Ins. / Ins. Pause / Exp. Pause VACV, PACV, VSIMV, PSIMV, / Nebulizer On,Off / 100 % O2 SPONT, tBi-Level, NIPPV, - Alarm silence ( 2 min ) / Alarm Reset AUTO , Apnea Backup - Freeze / Vent Setting / Alarm Setting Body Weight 3 ~ 150 kg - Exit / Key-lock / Accept Pressure 5 ~ 80 cmH2O Pressure Support 0 ~ 45 cmH2O Volume 10 ~ 2500 ml Input and Output Communication RS-232 : Baud rate 115200 bps Respiration Rate 2 ~ 120 BPM Setting Parameters I : E Ratio 4:1 ~1:120 Monitoring Parameter Inspiration Time 0.2 ~ 9.9 sec Alarm status & Event Message Plateau Time 0 ~ 9.9 sec Patient Name / Bed Number / Date & time TSUS 0.2 ~ 24 sec Wave Display PEEP 0 ~ 45 cmH2O High PEEP 5 ~ 45 cmH2O Low PEEP 0 ~ 45 cmH2O E-sense 10 ~ 60 % Fend 25 ~ 100 % Display parameter Pressure / Flow / Volume / P-V / F-V Trend Display 72 Hrs Tabluar & Graphic Trend. Pneumatic Performance Pressure : +/- 5% or +/- 1cmH2O Trigger Pressure 0.5 ~ 20 cmH2O Flow : +/- 10% @ 1bar Flow 0.5 ~ 20 LPM Volume : +/- 10% @ 1bar O2 21 ~ 100 % R-R interval : +/- 1% or 1 BPM Nebulizing time 10 ~ 180 Min Power Pneumatic Connector Inspiratory / Expiratory Limb Conn. AC Power Input 100 VAC ~ 230VAC 50/60Hz ISO 22mm Conical Male Power Consumption Air and Oxygen Inlet Internal Battery DISS Male / Female 25 < 90W 12VDC 7A Pb Battery MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL External Power Input VER 1.1 November, 2008 11VDC ~ 15VDC Internal Battery ** Optional : Additional battery available Dimensions / Weight Dimensions: 38 (W) x 43 (D) x 69 (H) (cm) Weight : 35 Kg Optional CART Dimensions : 49 (W) x 56 (D) x 77 (H) (cm) Weight : 28 Kg Environmental ℃ (50℃ ~ 104℃) Storage Temperature -10 ~ 70 ℃ (14℃ ~ 158℃) Operating Temperature 10 ~ 40 Relative Humidity 10 ~ 85 % (storage) 26 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 6. Description of each board (1) System Block Diagram (over all system) Block Diagram for Hardware UART OSC 16C554 SRAM TFT 10.4' LCD RS- 232 CENTRAL SYSTEM ARM FLASH S3C2410A LED DISPLAY VGA OUTPUT SDRAM LCD MONITOR DSC B/ D KEY FRONT B/ D IO PWR RS- 232 CONV SOUND ADAC SPEAKER DRIVER B/ D AC BOOST- UP PWR SV1 TMS320LF2406 RS- 232 RAM +24V 1.0A SV2 DSP POWER B/ D AD POWER +15V 4.0A ROM STEP- DOWN PWR +5V 1.0A MONITOR PWR +15V 1.0A SV3 IO PNEUMATIC B/ D PWR CONVERTER ADC SV4 PV1 PWM LS1 O2 SENSOR P1 DP1 P2 DP2 P4 DP4 DIFF AMP DIFF AMP DIFF AMP DIFF AMP DIFF AMP DIFF AMP PRE SEN PRE SEN PRE SEN PRE SEN PRE SEN PRE SEN 27 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (2) DSC Board A. DSC Board Picture B. Block Diagram ROM OSC STATIC MEMORY RTC & TIMER ARM S3C2410A SRAM FLASH Address Data SDRAM CONTROLLER KEY_BOARD CONTROLLER SDRAM JTAG RS232C VGA CONTROLL DAC UART 16C554 SOUND DAC 28 TFT LCD SPEAKER SOUND MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 C . Function Description 1) S3C2410X01 RISC MICROPROCESSOR 1.8V int., 1.8V/2.5V/3.3V memory, 3.3V external I/O microprocessor with 16KB ICache/16KB D-Cache/MMU External memory controller (SDRAM Control and Chip Select logic) LCD controller (up to 4K color STN and 256K color TFT) with 1-ch LCD-dedicated DMA 4-ch DMAs with external request pins 3-ch UART (IrDA1.0, 16-Byte Tx FIFO, and 16-Byte Rx FIFO) / 2-ch SPI 1-ch multi-master IIC-BUS/1-ch IIS-BUS controller SD Host interface version 1.0 & Multi-Media Card Protocol version 2.11 compatible Watch Dog Timer 117-bit general purpose I/O ports / 24-ch external interrupt source Power control: Normal, Slow, Idle and Power-off mode 8-ch 10-bit ADC and Touch screen interface RTC with calendar function On-chip clock generator with PLL 2) VGA INTERFACE 15Pin D-sub (CON5) LCD Display (receive the measured data) 3) RS-232 INTERFACE 9pin D-sub (COM1) RS-232 communication port purpose is Software upgrade. ( Program : ArmDown 3.1) 4) POWER ON/ OFF & KEY B/D INTERFACE 26Pin D-sub (CON13) From the Pneumatic it uses in Power supply and Data transmission. 29 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 D . DSC Software Description 1) DSC Software Architecture 2) DSC Software Architecture Description NO. PROCESS DESCRIPTION Process real-time based wave. 1 Wave Main Display base wave and waves in each mode. Process the data related to displaying the area of wave. 2 Vital Sign Main Process and display the data in each vent mode. Display the data processed by Pneumatic B/D. Adjust the amplitude of the wave received from Pneumatic B/D 3 Wave Gen. to fit it to the boundaries of to be displayed in LCD. Calculate the quantity of wave per unit time. 4 Vital Sign Gen. 5 COMM Main Calculate or Set interval between each Vital Sign received from Pneumatic B/D to make it be easily recognized by user. Transmit and receive data related to communication with Pneumatic B/D and External Serial Port. 30 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL 6 COMM Rx. 7 COMM Tx. VER 1.1 November, 2008 Classify Vital Sign, Wave Data, Instructions and status information received from Pneumatic B/D. Transmit user's setting value, Vent mode, Instructions, Information corresponding to alarms to Pneumatic B/D. Control time and interval needed to each task. Control time interval of calculation related to Wave and Vital 8 TIMER Main Sign display. Control interval of storing Trend data. Control time of warning related to alarm and interval of making a bundle. Execute instructions corresponding to Vent mode set by user and saved Set Value in each mode. 9 Invoke CMD Execute and Command instructions of needed behaviors corresponding to saved information of user such as Alarm, bed information, patient information and selection of language. 10 Menu Main 11 Menu Gen. Create and Display User Pop Up Menu and save Menu selected by user. Display Menu and Create and Remove next menu corresponding to selection in previous menu option. Manage to display whole screen. 12 Window Main Set the sequence of display. Display Wave, Vital Sign, Message, Alarms, Trend Data. 13 Window Gen. To create one full screen, make several sub-screen be being used and decide which sub-screen and area to be displayed. Manage the screen displayed last time and display Wave, Vital 14 Screen Main Sign and Window Screen making a decision considering the order of priority and displaying of screens or not. 31 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 3) DSC Software Algorithm Flow Chart 32 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 4) DSC Software Algorithm Description NO. PROCESS DESCRIPTION Execute initializing after Power On. 1 Vent Initial Display initial screen and store set information corresponding to initial set value. 2 Tx Initial Com. To Pneumatic 3 Initial OK 4 Vent Stop 5 User Command 6 User Information 7 Save User Setting 8 Change User Window 9 Vent Mode Transmit initial set value and command of initializing to Pneumatic Block. Receive the value of status of initialized pneumatic block. Display message and stop operation of ventilator when pneumatic block has problem in initializing. Receive user's command and execute needed operation. Receive user's information such as time, patient information, bed information, selected language etc. Store the information related to patient and set configured value. Manage to display in accordance with patient information. Receive the mode of ventilation by user. Receive the set value corresponding to ventilation mode 10 Input Vent Setting such as Pressure, Weight, Pneumatic flux, Pneumatic specific characters. 11 Auto Mode 12 Input Weight Receive set ventilation mode by user. Receive patient's weight information by user. Calculate not only value corresponding to user's set 13 Calculate Mode ventilation mode and set value in each mode but also value to be proper for communication. 14 Save Vent Setting 15 Change Vent Window 16 Tx Pneu B/D 17 Rx Pneu B/D 18 Save Pneu Setting Store set value and calculated value in each mode. Change window corresponding to user's ventilation mode and set value. Transmit set value and ventilation mode to pneumatic block. Receive information and instruction related to pneumatic block from it. Store the information and execute operation related to pneumatic block. 33 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 19 Vent Start 20 Do Ventilation 21 Message Display 22 User Interrupt Transmit ventilation mode executed by user to pneumatic block and ready to display. Display window and pneumatic information value in operation. Display all kinds of alarm and information message occurred in operation. Receive user's instruction in operation such as Power off, Mode changing, Set value changing. Transmit user's pause instruction to pneumatic block and 23 Vent Stop display message related to pause instruction and stop doing ventilation. 34 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (3) PNEUMATIC B/D A. PNEUMATIC Board Picture B. Block Diagram PNEUMATIC MAIN B/ D +5VS DC- DC Convertor +24V +V_POWE R +5V RS- 232C 19200,8,N,1 +3.3V CPU TMS320 LF2406 BATT SEN POWER ON/ OFF CONTROL FAIL CIRCUIT LCD POWER Control +24V DIFFERENCE CIRCUIT INOP SIGNAL WATCH DOG LATCH CIRCUIT LATCH CIRCUIT CONTROL LOGIC OVP CIRCUIT AD CONVERTOR PWM CIRCUIT DSC Board COMMUNICATION SV1 CONTROL SV1 SV2 CONTROL SV2 PV1 CONTROL PV1 SV3 CONTROL SP3 SV4 CONTROL SP4 LS1 CONTROL LS1 O2 Sensor DP 1 +5V DIFFERENCE CIRCUIT +3.3 V DIFFERENCE CIRCUIT +3.3 V INLET FLOW DP 2 P1 +5V +5V DIFFERENCE CIRCUIT +3.3 V INLET PRESSURE DP 4 P2 +5V DIFFERENCE CIRCUIT +3.3 V 35 DIFFERENCE CIRCUIT +3.3 V INSPIRATION PRESSURE INSPIRATION FLOW +5V P4 +5V DIFFERENCE CIRCUIT +3.3 V EXPIRATION FLOW EXPIRATION PRESSURE MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 C . Function Description 1) TMS320LF2406A DSP CONTROLLER High-Performance Static CMOS Technology − 25-ns Instruction Cycle Time (40 MHz) − 40-MIPS Performance − Low-Power 3.3-V Design On-Chip Memory Boot ROM (LF240xA Devices) Up to Two Event-Manager (EV) Modules (EVA and EVB), Watchdog (WD) Timer Module 10-Bit Analog-to-Digital Converter (ADC) Serial Communications Interface (SCI) 16-Bit Serial Peripheral Interface (SPI) Up to 40 Individually Programmable, Multiplexed General-Purpose Input/Output(GPIO) Pins Up to Five External Interrupts (Power Drive Protection, Reset, Two Maskable Interrupts) Real-Time JTAG-Compliant Scan-Based Emulation, IEEE Standard 1149.1† (JTAG) 2) DRIVER B/D INTERFACE From the DRIVER B/D it receives the Data which is necessary to a POWER supply and a Sol Valve control. From the CPU (TMS320LF2406A) it transmits the Data which controls Sol Valve with DRIVER B/D. 3) SENSOR INTERFACE Connect the Header Pin between the Sensor Board and Pneumatic Board Is used in data transmission. 4) RS-232 INTERFACE 9pin D-sub (COM1) Software upgrade. (Write240x. Ver 1.06) 36 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 The RS-232 port of the MV2000 is used the Debugging port. (Ventilator DSC Simulator Ver 1.0) 5) POWER ON/ OFF & KEY B/D INTERFACE 26Pin D-sub (CON13) The Serial port of the MV2000 is used in Power supply and Data transmission D . Pneumatic Software Description 1) Pneumatic Software Architecture 2) Pneumatic Software Architecture Description NO. PROCESS DESCRIPTION Store and process values sensed by each sensing part. DP1 : Incoming flow flux value into tank 1 Sensor DP2 : Inspiratory flow flux value DP4 : Expiratory flow flux value P1 : Pressure value in tank. 37 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 P2 : inspiratory pressure value P4 : expiratory pressure value, O2 Sensor : Oxygen Concentration. Battery Sensing : remaining battery charge 2 A/D Converter 3 Filtering Adjust received data from each sensor for fitting to the range of calculation. Extract original signal filtering noise from signal sensed by sensors. Calculate inspiratory and expiratory pneumatic component to 4 Calculation and from patient through figures of flux, pressure and concentration of flow. Check exhausting of wanted pneumatic component and 5 Processing Result monitor the status that equipment operates smoothly without malfunction. Control flux, pressure and blending ration of flow incoming 6 Result into pneumatic tank corresponding to processed result and display status message. 7 DSC Command 8 Module Command 9 COMM Transmit messages such as Alarm, Status, Command corresponding to result to DSC Process the condition of blending and instruction based on the weight and condition received from DSC. Manage communication between DSC and PNEU B/D. 38 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 3) Pneumatic Software Algorithm Flow Chart 39 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 4) Pneumatic Software Algorithm Description NO. PROCESS DESCRIPTION 1 Vent Initial 2 Initial OK 3 Vent Error 4 Vent Status Tx 5 Vent Mode Status Rx 6 Vent Start Comm. 7 Vent Mode Calculation 8 Vent Mode Start 9 Do Ventilation Ventilate gas and control valves in accordance with set value. 10 Mode Change Check change of set value by user. 11 Input Sensor Output 12 Pneu. Block Status Tx 13 Control Pneu. Block 14 Stop Vent. Receive initial instruction and set value from DSC B/D after Power ON. Check Sensors and Valves that are on pneumatic block. Store information related to malfunctions of sensors and valves. Transmit information of status of pneumatic block like normal operation, and malfunction to DSC B/D. Receive user's set value and pneumatic information from DSC B/D. Wait user's instruction of execution automatic and vent mode from DSC B/D. Calculate current mode set value after receiving start command from DSC B/D. Operate ventilator Receive values related to pressure, flow flux, oxygen concentration from sensor. Transmit information of current status of pneumatic block to DSC B/D. Control gas blending ratio, pressure, flow flux, specific characters comparing value from sensor with configured value. Stop operation of pneumatic block receiving the command of stop operation of ventilation. 40 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (4) DRIVER B/D A. DSC Board Picture B. Block Diagram BOOST-UP POWER Sol Valve CONTROLL POWER +24V / 1A SV1 DRIVER CONTROLL POWER SV3 SV2 +24V / 1A INPUT (+15V/4A) +5V / 0.5A SV4 STEP-DOWN POWER PV1 DRIVER PV1 +5V / 1A LS1 DRIVER LS1 PNEUMATIC BOARD CONNECTOR 41 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 C . Function Description DRIVER B/ D is consisted of POWER Connection part, Sol Valve Connection part (SV1, SV2, SV3, SV4, PV1, LS1), Pneumatic Connection part (POWER ON/ OFF & KBD INTERFACE), DC/ DC Power GENERATOR part, DRIVER part. Because POWER B/ D Connection part supplies DC voltage to DRIVER B/ D and Inputed DC voltage becomes DC/ DC POWER GENERATOR part relationship input, display DC voltage (+24V) that need in solenoid Valve control. These voltage are used to Sol Valve control that is linked to Sol Valve connection part. Driver part is inputed PWM signal that need PV1 and LS1 Valve control by Pneumatic connection part (POWER ON/ OFF & KBD INTERFACE) and Flow does control. 1) POWER B/D INTERFACE Supplied DC +15V in Power Board 2) D/D POWER GENERATOR Display +24V power who need in Sol Valve control. 3) PV1, LS1 DRIVER Compare Reference Voltage with PWM Signal that is displayed from Pneumatic Board and control PV1 valve and LS1 Valve. 4) POWER ON/ OFF & KEY B/D INTERFACE Header pin(2*14) Power supply and Data transmission 42 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (5) SENSOR B/D A. DSC Board Picture B. Block Diagram Pressure Sensor Sensor OUT + - Detection +Vref Sensor B/D Pneumatic 43 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 C . Function Description SENSOR B/ D is consisted of comparator. Input of comparator is Reference Voltage ((- ) input) and being Pressure SENSOR's output voltage (( +) input) compare two inputs and detect Pressure. Each Sensor B/ D Sol Valve according to design value because do Open, Close Pressure regulate. [Reference] • MV2000's OVER Pressure 3 steps design structure ①. Mechanical OVER Pressure : Design structure that is considered by last method after biographical drive Fail generation. ②. Hardware OVER Pressure : Design structure by comparison of immobile Hardware set point. ③. Software OVER Pressure : - User is OVER Pressure design structure that change is possible - Establish pressure automatically safety via maximum pressure in action state. 44 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (6) KEY B/D A. KEY Board Picture B. Block Diagram DSC B/D Speaker Connector DSC INTERFACE RED_LED RED_LED Connector ENCODER GREEN_LED GREEN_LED Connector Push Button C . Function Description KEY Board is consisted of DSC Board connector part, Speaker connector part, LED conn ection part. DSC Board connector part does role that supply DC voltage (+5V) to KEY Board and role that connect input Signal by DSC Board and Pneumatic Board in Push Button. Because Speaker connection part and LED connector part are linked with Speaker and A larm LED, if problem happens to product, Beep sound and Alarm Lamp happen. 45 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (7) POWER B/D(AC-DC) A. POWER Board Picture B. Block Diagram POWER A/D INPUT(AC) ONPUT(DC) TRANSFORMER EMI/NOISE FILTER INPUT RECTIFIER BATTERY CHAGER O.S.C CIRCUIT SWITCHING CIRCUIT +15V OUTPUT CURRENT LIMIT BATT +15V HEATER CIRCUIT FEED BACK CONTROLL 46 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 C . Function Description The first and the second of Power Board are insulated by Transformer. Also, EMI and Noise it is to do filtering EMI/ NOISE FILTER Part and there is Power supply part, Battery Charge part to supply DC voltage to the second. In case it is System Fail part who go Heater Curcuit part to run Heater beside him, and produce System FAIL is Beep sound and power of product does not become supply normally, Beep sound happens. a. General Characteristic This power supplying device is a converter with FLY BACK application which receives AC100V~ AC240Vand supplies DC+15V necessary for MV2000. AC input is FREE VOLTAGE which generates within the range of AC100V~ AC240V without external manual control. b. INPUT VOLTAGE AND FREQUENCY MV2000 A/ D Rated Output Voltage Input Variation Rate Frequency Range Remarks 100V~ 240V 45~ 66Hz FREE VOLTAGE AC 110V/ 220V single Phase MV2000 AC- DC CONVERTOR Rated Output Voltage Input Variation Rate Frequency Range Remarks DC +15V DC +15V (± 5%) - - c. Output Characteristics MV2000 A/ D NO. Description 1 Rated Output Voltage No- Load Voltage Range(%) Load Rated Current Load Voltage Range RIPPLE & NOISE (Rated Input & Rated Current) Over Current Protection (In Case of Short 2 3 4 5 6 +15V MAIN +15V +13.5V CHARGER +12.5V ± 5% ± 1% 2A ± 5% 1A +10V over 200mV P- P - AC100V- 240V, Above 6.3A Above 1.5A Below 5V 47 Remarks Time consumed for Battery recharge 6H± 30m MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 Output) 7 Output Power Battery Voltage above +12.5V 30W 12.5W * Input Voltage based on AC220V d. input Voltage Variation Rate In Case the rated current is loaded, even if the input voltage fluctuates as b, # 4 on C is satisfied the range of voltage. e. Load Variation Rate AC110V or 220V under rated input condition to satisfy the range of output voltage in case of # 3, 4 on C under loaded current. f. Efficiency (AC220 Input) Efficiency exceeds 65% for the measurement under MV2000 A/ D rated input and rated loaded current. g. Leakage Current In case of rated input condition on MV2000 A/ D, the leakage current between input LINE and F.G is below 200uA under AC 240V h. Inrush Current In Case of rated input condition on MV2000 A/ D, the inrush current is below 30A PEAK when the power is switched ON. 48 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 7. Assembly & Disassembly 1) Product Disassembly No Picture Description There is a LOCKER on the left side view. Please pull it out backward original from position its to separate main body from the CART. CART 1 Please lift up main body from the cart with care. And put it something on the flat like table to disassemble inside parts from the main body. 2 49 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 For separation of internal body from outer case, please loose four screws on the upper rear cover. 3 And loose 4 screws on the lower rear cover that is for BATTERY, BATTERY too. 4 Pull out internal body from outer case gripping those. 5 50 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 This is the photo shows the status of separation with each other. 6 To remove the POWER BOARD Please loose 3 screws on the plate covering it. 7 After removing screws, open the plate. 8 51 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 O2 SENSOR SENSOR Exchanging 1. Disconnect the harness for O2 SENSOR before exchanging it. 9 O2 SENSOR Exchanging 2. After disconnecting the harness from O2 SENSOR, please turn the O2 SENSOR CCW. 10 O2 SENSOR Exchanging 3 Lift it up after turning like left photo. And replace it with new one. When you try to exchange 11 only O2 SENSOR, you need to follow 9, 10, 11 instructions described above the opening door showed in the figure in chap.3-4 in this manual. 52 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 To separate the rear cover from the internal body, please turn trap CCW. * CAUTION * 12 Please breaking be careful the not plastic case. Like left photo, please loose the screw for TANK GAS releasing with hand. 13 This is the photo showing that the screw is loosed. 14 53 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 After loosing the screw, Please loose 7 screws fastening the rear cover. 15 The position of 7 screws is indicated in the left photo. 16 This is the photo separating the rear cover from the internal body after removing screws. 17 54 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 This photo shows the inside of internal body. 18 Please loose 2 screws indicated on the top of POWER SWITCH PANEL. 19 After removing screws, please lift up the panel and disconnect harness all connected of into POWER BOARD. 20 55 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 This photo shows the status of separated POWER SWITCH PANEL PANEL from internal body. 21 These two photos show the process removing the hose inserted onto NEBULIZER PORT. 22 These two photos show the process removing the hoses inserted onto PEEP MASS. 23 56 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 This photo shows the PEE PEEP VALVE LS1 after removing hoses inserted onto PEEP MASS to disconnect the harness connected to it. 24 These two photos show the process disconnecting the harness from PEEP VALVE LS1. 25 To pull out GAS TANK from internal body, please loose 4 screws are on the each side. 26 57 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 After removing 8 screws, please pull out the GAS TANK from internal body gripping like photo. 27 This photo shows the separated status of GAS TANK from internal body. 28 58 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 DRIVER Board Separation Description * DRIVER B/ D’ s Separation STEP 1 Picture is that separate in ternal block. I f s e p a r a t e D RI VER B oard, must remove SCRE 1 W. * DRIVER B/ D’ s Separation STEP 2 Separate DRIVER Board vertically. 2 59 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 SENSOR Board Separation Description * Remove SCREW of mar ked part. 3 Separate SENSOR Board vertically. * Warning 4 Board must separate certainly vertically. SENSOR B/ D is detached like picture. 5 60 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 PNEUMATIC Board Separation Description Remove SCREW and SUP PORT of marked part. SCREW : 2EA (SCREW of upper part) SUPPORT : 4EA (DRIVER Board fixing) 6 Board is detac hed like pi cture * Warning PNEUMATIC B/ D' s separa tion is possible though S ENSOR B/ D is detached. 7 61 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 Solenoid Valve Separation Description * Marked Sol Valve's name SV4 SV3 on each part of picture. * Reference When assemble, must co PV1 nnect jumper certainly. 8 Connect to relevant Conn ection Part of DRIVER SV2 SV1 Board using jumper that is offered in MEKICS. * Each Valve' s separation is as following. Remove nut using spanner 9 . Separate Valve Body like picture. If separate fixed Valve, m ust remove Screw. * Warning 1 2 Remove two Screws of di agonal line direction. direction. Befo 10 re remove remainder two screws, must remove all r emaining AIR to TANK. TA 2 1 NK's AIR observes becaus e is high tension. 62 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 * SV4 Valve If separate fixed Valve, must remove Screw. 11 Left side picture is that remove Valve. Second picture is SV4. 12 63 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 2) Monitor's separation No Monitor Separation Description Monitor Disassembly Please loose 4 screws fastened the MONITOR onto PRODUCT. 1 * CAUTION When separating the MONITOR, please grip it using one hand not to drop it down after removing the screws. This photos shows the separated status of MONITOR from PRODUCT. 2 64 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 After separating the MONITOR, please loose 4 screws indicated in the photo to remove the rear cover. 3 * CAUTION Please be careful not to make damage on the surface of TFT-LCD. After removing screws, lift up the rear cover. Upper side shows the SPEAKER. Lower 스피커 is composed of several parts. ⓐINVERTER for LCD ⓒ 4 side ⓑ DSC BOARD ⓒ TFTTFT-LCD PANEL ⓓ ⓓ LED ⓑ ⓐ ⓔ KEY BOARD ⓔ 65 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 MONITOR MAIN FRAME Separation Description To separate cover, rear please disconnect the wire connected onto KEY BOARD. 5 Next, disconnect two wires connected onto KEY BOARD that comes from LEDs which indicate the status of operation. 6 LED7(Green) NORMAL. indicates and LED6(Red) indicate ALARM. These two photos show the process to disconnect wires step by step. 7 66 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 And then, disconnect two wires for inverter connection. Left 8 photos show disconnection of upper wire. Disconnection of lower is wire identical with upper wire. The onto wire connected DSC BOARD is need to be disconnected from DSC BOARD. 9 These two photos show the process to disconnect the flat cable connected onto DSC BOARD. 10 * CAUTION The connector of flat cable is very weak so that need to be careful when it be opened. 67 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 After removing all harness into inside board, then loose 4 screws fastened the MAIN FRAME. The 11 position indicated in is the photo. The cable connected between DSC BOARD and KEY BOARD need to be disconnected. So, please lift up MAIN the 12 FRAME gripping like photo and disconnect the cable. These two photos show the process to disconnect the cable. 13 68 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 This photo shows the separated status of MAIN FRMAE from the outer case. 14 INVERTER BOARD Separation Description The INVERTER for LCD can be separated from the MAIN FRAME loosing 2 screws like left photos. 15 The INVERTER separated is displayed. 16 69 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 DSC Board Separation Description The DSC BOARD need to be separated from MAIN the FRAME loosing 5 screws are indicated in the photo. * TIP 17 DSC BOARD separated can from be the status of MAIN FRAME attached on outer case of MONITOR. This photo shows the separated status of DSC BOARD and FRAME. 18 70 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 TFT-LCD PANEL Separation Description P l e a s e , l o o se 4 screws TFTTFT-LCD to separate from MAIN FRAME. 19 This photo shows the separated status of TFTTFT-LCD PANEL and FRAME. 20 71 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 KEY BOARD Separation Description KEY BOARD fastened screws is by 10 onto rear side of outer case of MONITOR and its knob 21 is connected also in front of KEY PAD part. First, please remove the knob from the encoder. And then, loose 10 screws are indicated in the photo on the KEY BOARD in rear side of outer case. 22 72 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ALARM and NORMAL_LED Separation Description To which separate indicate LEDs the status of operation, please, loose 2 hex head screws. 23 This photo shows the separated status of LEDs and FRAME. 24 73 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 8. Voltage Control method (1) Each Board's Voltage control method 1) Digital Board’s Voltage Control Sol valve Connector Port SV1 SV2 SV3 SV4 PV1 Pneumatic Connector Port PV1 CONTROL VR1 TP3 POWER Connector Port C34 LS1 CONTROL VR2 TP1 TP2 DRIVER B/D Picture LS1 Figure 1. DRIVER B/D ① Each B/ D(POWER, DRIVER, Pneumatic, DSC, Sensor) was linked uprightly. ② Supply AC power (220V/ 60Hz) to POWER B/ D. ③ Voltage (error range : ± 2V) of +15V inputed to DRIVER Board. (TEST Point : TP1) ④ Voltage of +24V are displayed to TEST POINT(TP2). (error range:±1V) ⑤ Voltage of +5V are displayed to TEST POINT(TP3). (error range:±0.5V) ⑥ Voltage of +5V are displayed to C34. (error range:±0.5V) This time, C34's measurement does in ( +) terminal. ※ Measurement uses digital multi- meter(DMM) that recognize in MEK. With upside( ① ~ ⑥), execute voltage control. 74 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 2). Sensor Board’s Voltage Control DP4 RS232 (serial) DRIVER Connector P4 Sensor Connection Port DSC Connector Port P2 TP15 TP14 DP2 Pneumatic B/D Picture O2_Sensor regulation VR14 +3.3V regulation VR15 Figure 2. Pneumatic B/D B/D ① Pneumatic Board’ s TEST POINT(TP15) Voltage : +3.3V (error range:± 0.01V) ※ Voltage control does necessarily because influence in product performance. ② P1 Sensor's TEST POINT (TP4) Voltage : (+) 1.012V (error range:± 0.02V) ③ DP1 Sensor's TEST POINT (TP4) Voltage : (+) 1.650V (error range:± 0.02V) ④ P2 Sensor's TEST POINT (TP4) Voltage : (+) 1.178V (error range:± 0.02V) ⑤ DP2 Sensor's TEST POINT (TP4) Voltage : (+) 1.650V (error range:± 0.02V) ⑥ P4 Sensor's TEST POINT (TP4) Voltage : (+) 1.178V (error range:± 0.02V) ⑦ DP4 Sensor's TEST POINT (TP4) Voltage : (+) 1.650V (error range:± 0.02V) ※ Voltage of each sensor control according to ②~ ⑦. 3) O2 Sensor's voltage control ① Reference Voltage adjustment (VR15) : Pneumatic Board’ s TEST POINT(TP15) Voltage : +3.3V (error range:± 0.01V) ※ regulate +3.3V and measure O2 Sensor's voltage. ② O2 Sensor Voltage adjustment (VR14) : Pneumatic Board’ s TEST POINT (TP14) Voltage : (+) 1.01V 75 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (2) Product inspection method - Complete voltage control. Do program download through Pneumatic Board's RS- 232 Port. Must do inspection such as lower part. (Use Up- Load Cable that offer in MEK) 1) Sol Valve on/off TEST Figure 3. DSC PROGRAM ① Execute after DSC PROGRAM establishment. - DSC.EXE (Practice file) ② Connect communication JIG to COM1 port of computer (Supply Air, and product power to do on) ③ Select COM1( ) in upside Port establishment part and ( ) press on. This time, if press PWR ON, Flow wave and Pressure Wave appear on left side screen. 76 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 ④ First, execute , and press (Upside Setting becomes . , and practice after about 10 seconds ) ⑤ Confirm that is marked as "0.0" to below Window. ⑥ Inspect Valve's(SV1, SV2, SV3, SV4) on/ off action. (After execute : ①~⑤) Solenoid Valve “ON”: “1”, Solenoid Valve “OFF” : “0” ⑦ Each Valve(SV1, SV2, SV3, SV4) does on/ off TEST. 2) PV1 Adjustment ① Input PWM value to ⓐ. This time, PWM value confirms input value by Pneumatic Software Version. (ex. Ver0.80의 PWM value : 3603 [It is 10L/ m] ) a ② ⓑ To whether is displayed by 10L/ m confirm. If there is error, control DRIVER Board's VR1(Volume Resistor) by 10L/ m. b ③ Process of ①~② 3th repeatedly velocity of flow(L/m) confirm. 77 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL ※ 10L/ m ( VER 1.1 November, 2008 + within 1) of inhalation of air velocity of flow is kept to become. 3) LS1 Adjustment ① LS1 adjustment executes after PV1 TEST completion. ② Input 10cmH2O value to ⓒ. This time, 10cmH2O value confirms input value by Pneumatic Software Version. (ex. Ver0.80 의 PWM value : 1416 [ It is 10cmH2O] ) c ③ ⓓ To whether is displayed by 10cmH2O confirm. If there is error, control DRIVER Board's VR2(Volume Resistor) by 10cmH2O. d ④ Process of ②~③ 3th Exhaust velocity of flow (cmH2O) confirm. 78 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 9. TROUBLE SHOOTING NO BOARD Trouble Description Referred PAGE #1 DSC Board LCD screen badness 80 Solenoid Valve’ s Operation Badness 82 Unable to measure VENT MODE 83 KEY Operation Badness 84 #2 #3 DRIVER Board Pneumatic Board #4 KEY Board #5 POWER Board Power Deficiency & Unable to turn on the device 79 85 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (1) LCD Screen Badness LCD sc reen b ad ness Go o d Co nnec tio n b etween the Pneum atic B/ D and DSC B/ D Cab le? NO Is Cab le c o nnec ted c o rrec tly YES No rm al Op erate (DSC Cab le) NO YES Is Cab le in g o o d c o nd itio n? NO Rep lac e Cab le No rm al YES NO YES Sep arate m o nito r b ac ksid e Go o d Co nnec tio n Between the Bo ard Jum p er? NO Is Bo ard Jump er c o nnec ted c o rrec tly. (DSC B/ D, INVERTER B/ D, LCD' s Jump er) No rm al YES NO YES Is Jum p er in g o o d c o nd itio n? NO Rep lac e Jum p er (DSC B/ D, INVERTER B/ D, LCD' s Jump er) No rm al YES NO YES Go o d Co nnec tio n b etween the DSC B/ D KEY B/ D Jum p er? NO Is Bo ard Jump er c o nnec ted c o rrec tly. (DSC B/ D, KEY B/ D) No rm al YES NO YES A 80 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 A Is Jumper in good condition? NO Replace Jumper Normal (DSC B/ D, KEY B/ D) YES Operate NO YES A/S Request 81 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (2) Solenoid Valve’s Operation Badness Solenoid Valve's Operation badness Separate product backside Good Connection between the Power B/ D and Driver B/ D Jumper? NO Is Jumper connected correctly (Power B/ D, DRIVER B/ D) Normal YES Operate NO YES Is Jumper in good condition? NO Replace Jumper (Power B/ D, Driver B/ D Jumper) Normal YES NO YES Good Connection between the Driver B/ D and solenoid vlave Jumper? NO Is Jumper connected correctly (Driver B/ D, Solenoid Valve) YES NO YES Is Jumper in good condition? Normal NO Replace Jumper (Driver B/ D, Solenoid Valve's Jumper) Normal YES NO YES A/S Request 82 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (3) Unable to measure VENT MODE Unable to measure VENT MODE Good Connection between the Air and O2 Cable? NO Is Air and O2 Cable connected correctly NO Replace Cable (AIR, O2 Cable) Operate Normal YES NO YES INS and EXP Hose Connected correctly? YES NO YES Is Cable in good condition? Normal NO Is INS and EXP Hose connected correctly Normal YES NO YES Is Hose in good condition? NO Replace Jumper (INS, EXP's Hose) Normal YES NO YES A/S Request 83 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (4) KEY Operation Badness KEY Operation Badness Good Connection between the Pneumatic B/ D and DSC B/ D Cable? NO Is Cable connected correctly (Pneumatic, DSC B/ D) YES Operate NO YES Is Cable in good condition? Normal NO Replace Jumper ( Pneumatic, DSC B/ D Cable) Normal YES NO YES Separate product backside Good Connection between the DSC B/ D and KEY B/ D Jumper? NO Is Jumper connected correctly (DSC B/ D, KEY B/ D) Normal YES NO YES Is Jumper in good condition? NO Replace Jumper (DSC B/ D, KEY B/ D) Normal YES NO YES A/S Request 84 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 (5) Power Deficiency & Unable to turn on the device Power Deficiency & Unable to turn on the device Yes No Properly connect POWER CORD Check Power Cord Connection Normal Operation No Yes Yes No Replace Power CORD Is POWER CORD in good condition Normal No Yes Is the Fuse in good condition in the Fuse box? No Yes Replace Fuse(250V/ 6.3A) Normal No Yes Disassemble the rear cover of the device Separate cover for POWER MODULE Is there a good Connection between Power Cord and Power B/ D No Yes Connect Power Cord and Power B/ D properly Normal No Yes A 85 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 A Cable Connection between DSC B/ D and Power B/ D No Connection between DSC B/ D and Power B/ D Yes Normal No Yes Proper Cable Connection among each board Operation No Cable Connection among Each B/ D Yes Normal No Yes A/ S Request 86 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 10. Communications VGA 5 10 15 Pin Name 4 9 3 2 8 14 7 13 1 6 12 11 Direction Description ① RED Red Video (75ohm, 0.7p-p) ② GREEN Green Video((75ohm, 0.7p-p) ③ BLUE Blue Video(75ohm, 0.7p-p) ④ ID2 Not connected ⑤ GND Ground ⑥ RGND Red Ground ⑦ GGND Green Ground ⑧ BGND Blue Ground ⑨ KEY Not connected ⑩ SGND Sync Ground ⑪ ID0 Not connected ⑫ ID1 or SDA Not connected ⑬ HSYNC or CSYNC ⑭ VSYNC Vertical Sync ⑮ ID3 or SCL Not connected Horizontal Sync (or Composite Sync) Figure 99-1. Ventilator System VGA serial port pinout 87 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 11. Part List Part No. Part Name Description 140600020 Bacteria Filter Reusable 22OD-22ID 61-20-200 140601280 Bacteria Filter Spare Membrance 61-21-100-1 140601690 Breathing circuit set , Adult 140601700 Breathing circuit set ,Infant 140601410 Cable Monitor 140601660 Catheter Mount Tubings w 90 ˚ swivel 13cm Display Panel Cable 65-60-130 142401530 CBL H MV200-13 142401610 CBL H MV200-21 141200020 DIS TFT LCD 10.4 PD104VT1 141400030 High Pressure Hose, AIR 3.5m 141400020 High Pressure Hose , O2 3.5m 140601750 HME Spare Membrance 61-21-300-1 140600170 Humidi Chamber Adult MR370 140600880 Humidi Chamber Infant. MR340 13510702A MOD MV200 DSC REV.1A 13510901A MOD MV200 FRONT REV.1A 13511801A MOD MV200 DP1 REV.1A 13511811A MOD MV200 DP2 REV.1A 13511831A MOD MV200 DP4 REV.1A 13511841A MOD MV200 P1 REV.1A 13511851A MOD MV200 P2 REV.1A 13511871A MOD MV200 P4 REV.1A 140600431 MR810 Humidifier w/ 2 Chamber 140601340 MR850 Humidifier w/ 2 Chamber 140404040 MV2000 Cart 140601500 Nebulizer Kit 3561 141700380 OP Manual MV2000 MV2000 140100810 PEEP Valve(LS1) ACTUATOR 140601630 PEEP Valve Pad V20MPAD351A 88 Remark Adult/Pediatric (2 Chambers each) - MR810 Adult/Pediatric (2 Chambers each) - MR850 MEK Medical Equipment of KOREA MV2000 SERVICE MANUAL VER 1.1 November, 2008 140600530 Power Cable 141200020 PW INVERTER 140601680 Reusable HME 22OD-22ID 140500100 Sensor Oxygen IT DR M-04 140601650 Silicone Tubings ,adults 40cm 66-22-040tr 140601640 SiliconeTubings , adults 80cm 66-22-080tr 140601720 Silicone Tubings, Infant 45cm 66-11-045 140601710 Silicone Tubings , Infant 85cm 66-10-085 141000190 Solenoid Valve SV3 601100-A01 Burkert 141000200 Solenoid Valve SV1, SV2 6013SB-A02 Burkert 141000210 Solenoid Valve PV1 602200-A04 Burkert(2833) 141000410 Solenoid valve SV4 DSD-08AS DKC 140600950 Straight Conn.10OD-15OD 60-30-005 140600220 Straight Conn 22OD-22OD 60-12-110 140601510 Support Arm 140601740 Test Lung 500mL 6012 140601610 Test Lung 1000mL 6011 140600820 Water Trap, Adult 66-40-500 140601730 Water Trap, 10mmOD,Infant 66-40-510 140601670 Y-conn.straght15ID/22OD-22OD 60-10-100 140600210 90°Conn.22OD-22ID 60-12-000 140600940 90 AT-0151LS ˚Y-Conn. Infant, swivelling STD 89 60-20-100 MEK Medical Equipment of KOREA