BrainLAB IGRT System QA BrainLAB Frame-based System Target Localizer for Frame Based Treatment BrainLAB Frame-based System QA Winston Lutz Test Align Room Laser with Phantom Pointer Winston Lutz Test Gantry 0, Couch 0 Gantry 90, Couch 0 Gantry 180, Couch 0 Gantry 270, Couch 0 Gantry 0, Couch 270 Gantry 0, Couch 315 Gantry 0, Couch 45 Gantry 0, Couch 90 BRAINLAB FRAMELESS SRS/SRT WORKFLOW Infra Red System for Patient Pre-positioning Flat Panel X-ray for Image Fusion Patient Positioning and Treatment Image Registration and Calculation of Shift After Winston-Lutz Test Calibration of Isocenter Position ET Isocenter Phantom Verify the Isocenter Calibration • Place the E T isocenter phantom on the couch • Use automatic couch positioning to position the phantom • Make sure the laser align with the markings on the phantom X-ray Calibration • Place phantom on couch • Perform auto-positioning • Take x-ray image for each of the tubes • Verify that all 8 markers can be identified on the radiograph ET X-ray Calibration Phantom X-ray Calibration Validate the X-ray Calibration • Align ioscenter phantom with room laser • Take radiograph of the isocenter phantom • Perform image registration and couch shift • Check that the shift is within 1 mm Verifying Isocenter Position • Detect the Winston Lutz Pointer Center of the radioopaque sphere Isocenter determined during x-ray calibration