Quick Guide to using H-Cube® H-Cube® Components 1. Sample inlet line 2. Inlet pressure sensor 3. Gas check valve 4. Gas/liquid mixer 5. Bubble detector 6. Heater unit 7. CatCart® column unit 8. System pressure sensor 9. Back pressure regulator 10. Hydrogenated product collector 11. Touch-screen panel 11 5 10 4 1 2 9 8 6 7 3 H-Cube® Touch-Screen Interface 12. 13. 14. 15. 16. 17. Flow Rate Control (0,1 – 3 mL/min) Hydrogen Mode, see endnotes CatCart® Heater Pressure controller 1 bar = 14,5 PSI Stops the pump but keeps hydrogen Stops the pump and vents hydrogen If using elevated temperatures, always allow the system to cool to 40°C before attempting to remove the CatCart®. Ensure the water reservoir is topped up at all times. The cell must not be allowed to run dry or the cell will cease immediately and irreversibly to function. During use, the perspex cover on the H-Cube® must be closed. Spent CatCarts® should always be placed into deactivating vials prior to disposal. Normal COSHH stipulated procedures for the safe handling of chemicals should be followed at all times Fill up water tank Switch on H-Cube® Flow Solvent Through System Insert CatCart® Wait until you see solvent eluting from the reaction line!!! Set Temperature and Pressure and Mode on touch Screen, Press START CONTROLLED MODE FULL HYDROGEN MODE Wait until hydrogen is eluting from the reaction line and the hydrogen output is constantly under 25 in the bubble detector box on the Service Screen Wait until hydrogen comes out of the reaction line Transfer pump inlet line over to the reaction mixture When the reaction mixture is almost dry, add reaction solvent-Wash for 10 mins. When you have finished flushing the system with solvent press Stop or Stop and Keep H2 on the touchscreen and Stop on the pump • • • • For reactions at 1 bar-Use Full Hydrogen Mode For best results in controlled mode start at 50 bar For solvent compatability-see page 66 of the manual For information on system blockages-see p. 55 of the manual