GOM Acceptance Test 0000000476_001_EN_04-11-2014 Process Description Acceptance Test according to the Guideline VDI/VDE 2634 Part 3 GOM mbH Mittelweg 7-8 D-38106 Braunschweig Germany www.gom.com Tel.: +49 (0) 531 390 29 0 Fax: +49 (0) 531 390 29 15 E-Mail: info@gom.com Legal Notes Legal Notes No part of this publication may be reproduced in any form or by any means or used to make any derivative work (such as translations, transformations or adaptations) without the prior written permission of GOM. GOM reserves the right to revise this publication and change contents from time to time without obligation on the part of GOM to provide notification of such revision or change. GOM provides this manual without warranty of any kind, either implied or expressed, including, but not limited, to the implied warranties of merchantability and fitness for a particular purpose. GOM may improve or change the manual and/or the product(s) described herein at any time. Copyright © 2014 GOM mbH All rights reserved Page 2 (28) 0000000476_001_EN_04-11-2014 Table of Contents Table of Contents Legal Notes........................................... .....2 6 1 1.1 1.2 Important Notes.................................... .....5 Standard Signal Words................................5 Safety and Health Hazard Notes............ .....5 6.1 6.2 Restriction to the Operation Modes and Conditions.........................................17 Operation Modes.................................... ...17 Operation conditions..................................18 2 Introduction................................................7 7 Preparing the PSA................................ ...19 3 3.1 General Information about the Acceptance Test................................... .....9 Used Parameters.......................................10 8 8.1 8.2 4 Guideline VDI/VDE 2634 Part 3............ ...13 5 5.1 5.2 Differences between the Processes... ...15 Used Parameters.......................................15 Sensor Measurement Volume and Total Measuring Volume.................................. ...15 Test Positions.............................................15 8.2.1 8.2.2 8.3 8.4 8.5 8.6 8.7 Acceptance Test................................... ...21 Creating a New Project with VDI Test........21 Measuring Volumes with Differing Handling........................................................ ...22 Tilting the PSA........................................ ...22 Tilting by 30°........................................... ...23 Execute Measurement Series 1.................23 Execute Measurement Series 2.................26 Execute Measurement Series 3.................27 Evaluating Results.................................. ...27 Export Test Protocol...................................28 5.3 0000000476_001_EN_04-11-2014 Page 3 (28) Page 4 (28) 0000000476_001_EN_04-11-2014 Safety and Health Hazard Notes Important Notes 1 1.1 Important Notes Standard Signal Words In this publication the following standard signal words can be used: DANGER! ▶ The label points to an imminent danger. The situation can lead to serious bodily harm or death! WARNING! ▶ The label points to a dangerous situation. The situation can lead to serious bodily harm or death! CAUTION! ▶ The label points to a dangerous situation. The situation can lead to light bodily harm! NOTICE ▶ Info The label points to a situation which can lead to material damages. The damages can result on the product or in the vicinity of the product! The label indicates important application notes and useful information. 1.2 Safety and Health Hazard Notes CAUTION! Decalibrated sensor A decalibrated sensor leads to imprecise measuring result. ▶ Calibrate the sensor according to the respective hardware user manual! ▶ Maintain the warm-up time and the calibration limit values! CAUTION! Higher measuring uncertainty caused by mechanical vibrations Mechanical vibrations lead to higher measuring uncertainties. ▶ Measure only in environments which are free of vibrations! 0000000476_001_EN_04-11-2014 Page 5 (28) Safety and Health Hazard Notes Important Notes NOTICE Overexposed measuring areas In overexposed measuring areas, the sensor does not capture any measuring data. ▶ Avoid overexposures! ▶ Avoid extremely bright external light sources! NOTICE Higher measuring uncertainty caused by fluctuating temperature Fluctuating temperatures lead to higher measuring uncertainties. ▶ Keep the temperature constant throughout the measurement period! ▶ Page 6 (28) The ambient temperature and the temperature of the artifact have to be identical! 0000000476_001_EN_04-11-2014 Introduction 2 Introduction This manual is intended for qualified personnel who have comprehensive knowledge of coordinate measuring technology. Basic PC knowledge (windows-based programs and operating systems) is expected. For operating the system optimally, the ability to visualize in 3D and a color vision ability are assumed. 0000000476_001_EN_04-11-2014 Page 7 (28) Page 8 (28) 0000000476_001_EN_04-11-2014 General Information about the Acceptance Test 3 General Information about the Acceptance Test The acceptance test is performed in line with the GOM Acceptance Test according to the guideline VDI/VDE 2634 Part 3. The inspector creates the measured values for defined parameters using the process which is described in this process description. The inspector always performs the process in the same way. The inspector compares the measured values with the corresponding limit values. The limit values are determined by GOM and define the maximum permissible measurement error for the test. For the test, the inspector uses a sphere artifact which is produced by GOM. The sphere artifact with the name Probing / Spacing Artifact, short PSA (see Fig. 1), consists of several matching, homogeneous sphere pairs with different diameters and spacings. On behalf of GOM, the metric sizes (diameter of each sphere and centers of each sphere pair) are determined highly precisely in an external calibration. There are the calibration standards “manufacturer's calibration” and DAkkS calibration (DAkkS = Deutsche Akkreditierungsstelle GmbH is Germany's national accreditation body) available. Fig. 1: PSA400 (front) with 9 sphere pairs, PSA900 (back) with 4 sphere pairs If a sensor has several measuring volumes, the inspector tests each measuring volume in combination with the sensor in a separate acceptance test. In each acceptance test, the inspector measures one pair of spheres consisting of two spheres with the same nominal diameter size. The sphere pair to be used depends on the sensor to be tested and the used measuring volume. You find the assignment of the sphere pair to the sensor and measuring volume in the acceptance test protocol. In line with the acceptance test, the inspector measures three measurement series with ten individual measurements. The positioning and aligning of the sphere pair with respect to the sensor measuring volume is defined exactly for each individual measurement. The positioning and aligning not repeated within a complete acceptance test. The software carries out the transformation and fine-alignment of the individual measurements for each measurement series. From the measurements, the software computes a polygonized mesh. Subsequently, the inspector derives the values for the parameters. The inspector can export the results of the acceptance test as a PDF protocol. 0000000476_001_EN_04-11-2014 Page 9 (28) Used Parameters General Information about the Acceptance Test 3.1 Used Parameters To estimate the measuring accuracy, the software uses the following parameters: ● Probing error form (sigma) ● Probing error size ● Sphere spacing error ● Length measurement error To determine the corresponding current parameter values, the software calculates for a sphere pair the best-fit spheres from the measuring data. For the calculation, the software uses the method of least squares. The parameter probing error form (sigma) corresponds to the standard deviation. The standard deviation is calculated from the radial distance of all measured points of a sphere with respect to the corresponding fitting sphere with a freely estimated diameter (see Fig. 2). The parameter probing error size is the difference between the measured diameter and the calibrated diameter of the respective sphere (see Fig. 2). Fig. 2: Schematic representation of the calculation Probing error form Probing error size The parameter sphere spacing error is the resulting difference between the measured sphere spacing of both centers of the fitted spheres and the corresponding calibrated spacing (see Fig. 3). Fig. 3: Schematic representation of the sphere spacing error calculation The length measurement error is the difference between the measured length of two opposite points and their corresponding calibrated spacing. The measurement has to be carried out via a bidirectional probing from opposite direc- Page 10 (28) 0000000476_001_EN_04-11-2014 General Information about the Acceptance Test Used Parameters tions on nominal parallel surfaces perpendicular to one of the surfaces (see Fig. 4). Fig. 4: Schematic representation of the length measurement error calculation Info 0000000476_001_EN_04-11-2014 The ATOS measuring system measures full-field undifferentiated points. Due to this process, defined individual points cannot be determined specifically. Therefore, the single points which lie exactly on the connecting axis of both sphere center points and and point towards the outside on each sphere cannot be measured directly. That way, the parameter length measurement error can be determined only via an indirect combined measurement and evaluation. According to the guideline VDI/VDE 2634 - Part 3, the software determines the parameter length measurement error using Method C. Page 11 (28) Page 12 (28) 0000000476_001_EN_04-11-2014 Guideline VDI/VDE 2634 Part 3 4 Guideline VDI/VDE 2634 Part 3 The guideline VDI/VDE 2634 Part 3 with the title “Optical 3D measuring systems - Multiple view systems based on area scanning” contains a description on procedure and calculation for acceptance and reverification test of imaging systems with area-based sampling in several single scans. This group of systems includes especially fringe projection systems like the ATOS sensors which capture and evaluate an object using multiple individual measurements transformed into each other. The guideline was created by the committee of experts 3.32 “Optische 3DMeßtechnik” (Optical 3D Measuring Technique) and published in December 2008. The committee of experts 3.32 “Optische 3D-Meßtechnik” is a joint committee of VDI (Verein Deutscher Ingenieure = Association of German Engineers) and DGPF (Deutsche Gesellschaft für Photogrammetrie und Fernerkundung e.V., Arbeitskreis Nahbereichsphotogrammetrie = German Society for Photogrammetry, Remote Sensing and Geoinformation). 0000000476_001_EN_04-11-2014 Page 13 (28) Page 14 (28) 0000000476_001_EN_04-11-2014 Differences between the Processes 5 Test Positions Differences between the Processes GOM developed the GOM Acceptance Test from the guideline VDI/VDE 2634 - Part 3. The main purpose was the creation of a procedure which enables an exact and reliable statement with respect to the achievable measuring accuracy under realistic conditions. Furthermore, GOM kept the test as simple and stable as possible. As a result, there are several differences between both processes. 5.1 Used Parameters The acceptance test according to the guideline VDI/VDE 2634 Part 3 in line with the GOM Acceptance Test includes the following parameters: ● Probing error form (sigma) ● Probing error size ● Sphere spacing error ● Length measurement error The guideline VDI/VDE 2634 - Part 3 uses the same parameters. However, the probing error form is evaluated with the statistical size range. The result label and the protocol contain all results for both processes. That way, the test can be performed in compliance with the GOM Acceptance Test as well as with the guideline VDI/VDE 2634 Part 3. 5.2 Sensor Measurement Volume and Total Measuring Volume The guideline VDI/VDE 2634 Part 3 differs between the sensor measuring volume and the total measuring volume. The guideline VDI/VDE 2634 Part 3 understands the sensor measuring volume as the maximum measuring volume that is captured in one single scan. Usually, the sensor measuring volume is smaller as the total measuring volume which consists of all combined sensor measuring volumes of all single scans. According to the guideline VDI/VDE 2634 Part 3, the spatial diagonal of the total measuring volume is at least double as long as the one of the sensor measuring volume if nothing else was agreed on. Since the sensor measuring volume and total measuring volume are identical in the acceptance test in line with the GOM Acceptance Test, the correspondingly shorter distances are used for the normal. 5.3 Test Positions In line with the acceptance test according to the guideline VDI 2634 Part 3, the inspector places the artifact in different, defined places of the reference scene. The inspector measures the artifact with several single scans. In the process, the artifact is to be found in different positions with respect to the sensor measuring volume. For the probing error form and probing error size, the inspector measures at least three positions in the reference scene with at least five single scans each. For the sphere spacing error and length measurement error, the inspector measures seven defined positions. The number of single scans is not defined. In one single scan, the inspector measures only one of both test spheres if possible. The acceptance test in line with the GOM Acceptance Test differs in these points. The inspector tests the parameters in three measurement series with ten single scans each. Since the test spheres as well as the reference points are attached to the PSA base plate, the test spheres cannot be moved with 0000000476_001_EN_04-11-2014 Page 15 (28) Test Positions Differences between the Processes respect to the reference points. Almost always, the inspector measures both test spheres simultaneously. Page 16 (28) 0000000476_001_EN_04-11-2014 Restriction to the Operation Modes and Conditions 6 Operation Modes Restriction to the Operation Modes and Conditions The guideline VDI/VDE 2634 Part 3 recommends defining the operation modes and conditions for the acceptance test. Only under the defined requirements the corresponding limit values are authoritative for an acceptance test. According to the guideline VDI/VDE 2634 Part 3 the operation modes are the adjustment and configuration possibilities of the optical 3D measuring system. The operation conditions contain the environmental influences on the optical 3D measuring system. In line with the GOM Acceptance Test, GOM defines the following operation modes and conditions. The acceptance test of a system is performed according to this process description, observing the restrictions of the operation modes and conditions. The acceptance test protocol contains a reference to this process description. Due to that fact, the following operation modes and operation conditions are also part of the protocol. 6.1 ● ● ● ● ● ● ● 0000000476_001_EN_04-11-2014 Operation Modes The sensor and its parts are factory-adjusted. Check before a reverification measurement whether the settings comply with the specifications. If the settings do not comply with the specifications, set up the sensor according to the respective User Manual Hardware. This approach is not valid for ATOS Core sensors. Calibrate the sensor. Maintain the warm-up time and the calibration limit values. Carry out the measurements with the quality setting High and the resolution setting Full resolution. Choose the exposure time such that the measuring images are well exposed. Avoid overexposures. Polygonize the single scans to a mesh using the setting Standard. For calculating the spheres, the software uses only measurement data which is above a defined plane. This plane is aligned parallel to the artifact base plate. This plane intersects the sphere at the latitude 10° south. The software determines the spheres using the least squares method. In the process, the software rejects 0.3% of the measured values as outliers. This value corresponds to a setting of 3 sigma. The software determines the parameter Length measurement error using Method C (see VDI/VDE 2634 Part 3, page 18). Page 17 (28) Operation conditions Restriction to the Operation Modes and Conditions 6.2 ● ● ● ● ● ● ● ● ● Page 18 (28) Operation conditions The artifact, the PSA, includes two spheres fixed to each other with a base plate. The spheres have the nominal diameters defined in Operations ► Test VDI 2634 - Part 3 ► New... ► Artifact. The sphere spacing approximates the nominal distance defined in Operations ► Test VDI 2634 - Part 3 ► New... ► Artifact. The spheres are made from steel. To ensure diffuse reflection, coat the spheres with a titanium dioxide layer (see 7 Preparing the PSA). Sphere diameters and sphere spacing are determined by a DAkkS calibration. The corresponding measuring uncertainty has to be very small compared to the corresponding limit value. The PSA is equipped with reference point markers, cleaned and sprayed according to the instructions in ATOS User Manual- Probing Spacing Artifacts (PSA). Maintain the environmental conditions in accordance with the respective User Manual Hardware. In particular, the temperature must be kept constant throughout the measurement period. The ambient temperature and the temperature of the artifact have to be identical. The measuring environment must be free of mechanical vibrations. The ambient light must not vary extensively during the measurement. Avoid extremely bright external light sources. 0000000476_001_EN_04-11-2014 Preparing the PSA 7 Preparing the PSA Before you can use the PSA, e.g. for an acceptance test, equip the PSA with reference point markers. The reference points enable the transformation of the individual measurements into a common coordinate system within one measurement series. The size, the number and the distribution of the reference points directly influence the transformation quality and also the measuring results of the acceptance test. In ATOS User Manual- Probing Spacing Artifacts (PSA), you find detailed information regarding the handling of the PSA: ● Recommended point pattern for equipping the PSA with reference points ● Recommended sensor test configurations ● Handling of the spray stencil To be able to evaluate exactly the measuring accuracy of your system, prepare the surface with suitable means. Spray the PSA with titanium dioxide powder. In ATOS User Manual- Spraying with the Airbrush System, you find information and tips for handling the airbrush system purchased from GOM: ● Parts of the airbrush system ● General spraying process with the airbrush system 0000000476_001_EN_04-11-2014 Page 19 (28) Page 20 (28) 0000000476_001_EN_04-11-2014 Creating a New Project with VDI Test Acceptance Test 8 8.1 Acceptance Test Creating a New Project with VDI Test Requirements: ● ● ● ● ● Observe the safety and health hazard notes. Observe section 6.1 Operation Modes. Observe section 6.2 Operation conditions. Perform the acceptance test with an ATOS Professional software, minimum V8.0. The firmware has to be up to date. Procedure: 1. Check whether the prerequisites mentioned above are met. 2. Start an ATOS Professional software. 3. Import the PSA template with the respective nominal values. Info If you have purchased the PSA as of May 2013, you have to create manually the template using the corresponding calibration certificate (see ATOS User Manual- Probing Spacing Artifacts (PSA)). You have to create the template only once. a) Open Edit ► Application Settings ► Preferences... ► Templates ► VDI artifacts ► Import. Windows explorer opens. b) Navigate to the PSA template. Open the template. The software loads the template into the preferences. c) Finish the import with OK. 4. Create a project. a) Open File ► New Project. 5. Save the project under any name. a) Carry out File ► Save. The software saves the project in the chosen directory. 6. Create a VDI test. a) Open Operations ► Test VDI 2634 - Part 3 ► New.... b) Complete the dialog according to your requirements. c) Enter the measurement temperature. d) Enable ( ) the option Change number of test positions and set the value to 3. For the GOM Acceptance Test, three test positions (measurements) are required. e) Choose in Artifact the corresponding PSA template. 0000000476_001_EN_04-11-2014 Page 21 (28) Measuring Volumes with Differing Handling f) Acceptance Test Close the dialog with Create and close. The software creates three measurement series. The main toolbar of the workspace Digitize changes. You see additional buttons. Fig. 5: Additional buttons Info To set the active measurement series during the test, use only the buttons intended for this purpose which are described below. Using the function Switch To Previous Test Position, you navigate to the previous measurement series and set it as the active measurement series. Using the function Calculate Test Position, you start a calculation for the parameters of the active measurement series. This function is enabled as soon as you perform at least one measurement in the active measurement series. Using the function Switch To Next Test Position, you navigate to the next measurement series and set it as the active measurement series. Using the function Export Test Protocol, you export a PDF document containing all relevant measuring results of the GOM Acceptance Test. 8.2 Measuring Volumes with Differing Handling In line with the acceptance test, you measure three measurement series with ten individual measurements. The positioning and aligning of the sphere pair with respect to the sensor measuring volume is exactly defined for each individual measurement. In step 2 of the respective measurement series, you have to treat the following measuring volumes (see 8.3 Execute Measurement Series 1, 8.4 Execute Measurement Series 2, 8.5 Execute Measurement Series 3) differently from the standard process. 8.2.1 Tilting the PSA In the standard case, you tilt the sensor. During the process, the PSA remains in horizontal position in its case. With the following measuring volumes, you tilt the PSA and not the sensor in all three measurement series. Procedure: 1. Constantly hold the sensor perpendicular to the ground. 2. Tilt the PSA relatively to the sensor. Page 22 (28) ATOS Triple Scan II ATOS Triple Scan III ATOS Compact Scan 5M ATOS Compact Scan 2M ATOS Core MV 38 MV 38 MV 40 MV 35 MV 45 MV 60 MV 60 MV 70 MV 60 MV 80 MV 100 MV 100 - - - 0000000476_001_EN_04-11-2014 Execute Measurement Series 1 Acceptance Test 8.2.2 Tilting by 30° In the acceptance test, the tilting angle between the viewing direction of the sensor and the normal vector of the PSA is per default 45°. Nevertheless, use for the following measuring volumes the tilting angle of 30°. 8.3 ATOS Triple Scan II ATOS Triple Scan III ATOS Compact Scan 5M ATOS Compact Scan 2M ATOS Core MV 38 MV 38 MV 40 MV 35 MV 45 MV 60 MV 60 MV 70 MV 60 MV 80 Execute Measurement Series 1 Requirements: ● The project created in 8.1 Procedure: 1. Position the PSA in its case horizontally on a plane and stable surface. 2. Tilt the ATOS sensor by 45° with respect to its support axis (Fig. 6). Info If the sensor is attached to a lift or it is a SO measuring volume marked respectively by GOM (see 8.2.1 Tilting the PSA), tilt not the sensor but the PSA. Info For certain SO measuring volumes marked respectively by GOM, use the tilting angle of 30° instead of 45° (see 8.2.2 Tilting by 30°). Fig. 6: Positioning sensor and PSA for first measurement series (position for first measurement) Standard case 0000000476_001_EN_04-11-2014 Page 23 (28) Execute Measurement Series 1 3. 4. 5. 6. Acceptance Test Special case lift with respectively marked SO measuring volume Set the height by positioning the center of the measuring volume towards one of the spheres. Position the center of the measuring volume centrally towards the center point of the connecting axis between both spheres. Carry out eight measurements in this position. a) For each measurement, rotate the PSA by 45° around the set center point. b) Every time, position the center of the measuring volume centrally towards the center point of the connecting axis between both spheres. Position the PSA with the outside of the left sphere (outer piercing point of the connecting axis) into the center of the measuring volume (Fig. 7). Fig. 7: Positioning sensor and PSA for first measurement series (position for ninth measurement) Standard case 7. 8. 9. 10. 11. Special case lift with respectively marked SO measuring volume Carry out the ninth measurement. Rotate the PSA by 180°. Adjust the position of the PSA until the center of the measuring volume coincides with the piercing point of the connecting axis on the right outside of the sphere. Carry out the tenth measurement. Click button Calculate Test Position. The dialog Calculate Test Position opens. 12. Fill out the dialog. You can determine the spheres also with Ctrl + RMB. 13. Check the preview of the fitting plane and the point selection. 14. If the preview is satisfactory, confirm the calculation with OK. After a successful computation, the software displays the results in a result label. Page 24 (28) 0000000476_001_EN_04-11-2014 Acceptance Test Execute Measurement Series 1 Fig. 8: Result label for the first measurement series 0000000476_001_EN_04-11-2014 Page 25 (28) Execute Measurement Series 2 8.4 Acceptance Test Execute Measurement Series 2 Procedure: 1. Click button Switch To Next Test Position. The second measurement series is active. 2. Tilt the ATOS sensor from the perpendicular position by 45° in clockwise direction (Fig. 9). Info If the sensor is attached to a lift or it is a SO measuring volume marked respectively by GOM (see 8.2.1 Tilting the PSA), tilt not the sensor but the PSA. Info For certain SO measuring volumes marked respectively by GOM, use the tilting angle of 30° instead of 45° (see 8.2.2 Tilting by 30°). Fig. 9: Positioning sensor and PSA for second measurement series (position for first measurement) Standard case Page 26 (28) Special case lift with respectively marked SO measuring volume 3. From here, repeat the steps 3 to 14 of the section 8.3 Execute Measurement Series 1. 0000000476_001_EN_04-11-2014 Evaluating Results Acceptance Test 8.5 Execute Measurement Series 3 Procedure: 1. Click button Switch To Next Test Position. The third measurement series is active. 2. Tilt the ATOS sensor from the perpendicular position by 45° in counterclockwise direction (Fig. 10). Info If the sensor is attached to a lift or it is a SO measuring volume marked respectively by GOM (see 8.2.1 Tilting the PSA), tilt not the sensor but the PSA. Info For certain SO measuring volumes marked respectively by GOM, use the tilting angle of 30° instead of 45° (see 8.2.2 Tilting by 30°). Fig. 10: Positioning sensor and PSA for third measurement series (position for first measurement) Standard case Special case lift with respectively marked SO measuring volume 3. From here, repeat the steps 3 to 14 of the section 8.3 Execute Measurement Series 1. 8.6 Evaluating Results If all measurement series are evaluated successfully, the results for the parameters are determined. In the column Result, you see each largest deviation of the individual results of a parameter (Fig. 11). 0000000476_001_EN_04-11-2014 Page 27 (28) Export Test Protocol Acceptance Test Fig. 11: Result label for all measurement series If in all three measurement series the sensor meets all limit values for the parameters viewed, the acceptance test is successful. 8.7 Export Test Protocol Using the function Export Test Protocol, you export a PDF document containing all relevant measuring results of the GOM Acceptance Test. Page 28 (28) 0000000476_001_EN_04-11-2014