AL-106 and AL-106SA
Digital Video Colposcope
User Manual English Version 3.0
MedGyn AL-106 Digital Video Colposcope User Manual
Table of Content
1. Introduction
2
1.1 Scope
2
1.2 Indication
2
1.3 Features
3
1.4 System Introduction
3
1.4.1Main parts
3
1.5 Installation preparation
5
1.5.1Environment
5
1.5.2Unpacking and Inspection
6
1.6 Further information
6
2. Safety
7
2.1 Markings
7
2.2 Classification
7
2.3 Support and maintenance
8
2.3.1Warranty
8
2.4 Terminology
8
2.5 Safety Note
9
2.5.1Warnings
9
2.5.2Precautions
9
2.6 Contraindication
10
3. Camera
11
3.1 Composition
11
3.2 Installation
11
3.3 Disassembly
12
4. Stand
13
4.1 Vertical Stand
13
4.1.1Vertical Stand Composition
13
4.1.2Assembly
13
4.2 Swing arm stand
17
4.2.1Swing arm stand composition
17
4.2.2Assembly
17
4.3 Precautions
25
5. Camera Operation
26
5.1 Button Function
26
5.2 Camera Adjustment
28
1
I
MedGyn AL-106 Digital Video Colposcope User Manual
1. Introduction
Welcome to use AL- 106 digital video colposcope!
Clinical application of colposcope has a history of more than eighty years. With the
rapid development of medical diagnosis technology and digital imaging technology,
the design and function of the colposcope is becoming better day by day. The new
type of colposcope achieves much improvement in the advancement of the imaging
module, image clarity and personalized design.
AL-106 digital video colposcope adopts high - resolution electronic imaging technology.
This technique is different from traditional optical colposcopy for it does not have
eyepieces so that the doctors could observe the lesion on the monitor. The high quality
light source and continuously magnified image could present very subtle details.
In addition, AL - 106 is easy to operate, and can be used both for diagnosis and guiding
treatment. Such innovative and non - invasive technique could provide fast and safe
screening, and it is an advanced method for education and research.
1.1
Scope
Cervix, vulva and vaginal examination, treatment and data management
1.2
Indication
MedGyn's Digital Video Colposcope is intended for magnified viewing of the tissues
of the vagina, cervix and external genitalia in order to assist doctors in diagnosing
abnormalities such as lesions or cancer, and selecting areas for biopsy. The images
from the Digital Video Colposcope are to be viewed on a color monitor. The Digital
Video Colpscope is intended for use in hospitals, clinics, and doctors' offices.
2
MedGyn AL-106 Digital Video Colposcope User Manual
1.3
Features

AL-106 digital video colposcope adopts high resolution color CCD video image
capture technology and advanced control technology, thus can not only satisfy big
- screen display requirement, but also realize image ca pture, display, process,
save and print through computer and software.

High-resolution color CCD, double circular LED group light source, control
circuit all integrated in one handheld camera.

Button control provides convenience for operation such as zo om in/out, focus,
freeze, green filter, and timing and magnification factor display.

Digital video colposcope check is dual-visual that both doctors and patients
can observe, thus it could help doctor explain cervix lesion and suggest
treatment to patient directly.
1.4
System Introduction
AL-106 digital video colposcope is composed of camera and stand.
Composition and type
Diagram 1 -1 composition
Colposcope
Model
AL-106
AL-106SA
Camera
Conversion box
Vertical stand
Camera
Conversion box
Swing arm stand
1.4.1 Main parts
Camera
Camera is composed of high- resolution color CCD , circular LED group light source,
control circuit, it mainly used to control zoom in, zoom out, green filter, timing and
freeze function.
3
MedGyn AL-106 Digital Video Colposcope User Manual
Diagram 1 -2 AL-106 Camera
name
Camera
Description
1/4" SONY Exview IIAD CCD imaging module
Fast auto focus
magnification: 1~40
optical magnification: 1~36
double circular LED group light source
3 grade green filter function
S-Video output
Image freeze control
Working distance: 200~300mm
Field of view:
At 1X≥Φ60mm;
At 40 X≥Φ6mm
Depth of view: 5mm ~ 120mm
Magnification and acetic acid reaction timing
display
Multifunction board (direct control magnification,
focus, freeze and green filter))
Stand
Vertical and swing arm stand is used to support the camera and connect with the video
cable
Diagram 1 -3 AL-106 Stand
name
Vertical stand
Swing arm stand
Description
S-Video interface
Maximum height:≤900 mm
adjustable:≥150mm
movable 5 leg caster (with brake)
Vertical swing arm rotation:>180° ;
Left to right rotation :0-180°;
Up and down rotation:±30° ;
Stand height:800mm
Movable five legs base (three brake)
4
MedGyn AL-106 Digital Video Colposcope User Manual
Conversion box
Name
description
Conversion box
1.5
S-Video output
;
Signal interface (camera
out/in)
90~240V Power in
Installation preparation
1.5.1 Environment
To ensure electric assembly safety, the environment shall be reasonably dust free, no
corrosive or combustible gas, or extreme temperatur e or humidity.
Keep a proper space for the monitor at least 5cm from the wall to ensure good air
ventilation.
Keep the room clean, avoid shock, noise interference, and avoid direct sun.
Extreme temperature can affect the accuracy of the monitor and damage accessory or
circuit.
Please ensure that there is no water condensing in the device before use. When the device
is transferred between buildings, water condensing happens because of exposure to
humidity and temperature difference.
Operating room should be ventilated, anti -shock, noise, strong power and avoid direct
sunlight.
To ensure safe use, user must provide standard three- prong power socket and ensure
proper connection of earth protection especially when used together with microwave
machine or high frequency device.
5
MedGyn AL-106 Digital Video Colposcope User Manual
1.5.2 Unpacking and Inspection
Do not bottom up when unpacking. After unpacking the camera and accessories,
keep the packing materials properly for return service when necessary.
Check carefully the device and the accessories against the packing list supplied by
the manufacturer and see if there is any damage. Any damage or missed please
contact the agent or manufacturer.
1.6
Further information
This manual explains the performance, operation and maintenance of the colposcope.
If schematic diagram, key component list or any other materials needed by technical
personnel, please contact the manufacturer.
Note:
This manual should be put around the device for user. Read the manual
thoroughly before use to ensure proper operation, and safety.
6
MedGyn AL-106 Digital Video Colposcope User Manual
2. Safety
2.1 Markings
mark
description
mark
Note: check
accompany files
Fuse
description
○
disconnect power
│
connect power
AC current
USB interface
Signal input/output
Electronics pollution
control mark
high voltage
Outer packing is recyclable
recycle
Maximum stacking
Protect against
water
Fragile contents
Manufacturing date
up
Footswitch enclosure IP
code
series number
2.2 Classification
Classification information is as flowed:
Classification principle
According to the type of protection against
electric shock
According to the degree of protection against
electric shock
Classification
Class I
Type B
According to the mode of operati on
continuous operation
According to t he degree of protection against
Normal device, but footswitch is
ingress of water
IP68。
7
MedGyn AL-106 Digital Video Colposcope User Manual
According to the degree of safety in the
presence of a flammable anesthetic mixture
Non AP or APG equipment
with air or with oxygen or nitrous oxide
According to the methods of sterilization or
disinfection recommended by the
Recommend by manufacturer
manufacturer
According the type of management
Type Ⅱ,classification code 6822
2.3 Support and maintenance
2.3.1 Warranty
We provide warranty for this product within warranty period. And please keep the packages
for further transportation and storage.
We will assume the following responsibilities:

The device is assembled, adjusted, improved by authorized personnel.

Operation environment is in accordance with the electrical standard
specified in this manual.

Operate strictly according to this manual.
The warranty shall be assumed except:

Damage due to improper operation, improper connection to other device,
and accident.

The device is modified by user without the written authorization of the
manufacturer.

The serial number has been removed or unrecognizable.
2.4 Terminology
The terms danger, warning, and caution are used throughout this manual to point out
hazards and to designate a degree or level of seriousness. Familiarize yourself with their
definitions and significance.
8
MedGyn AL-106 Digital Video Colposcope User Manual
Note :
Provide application tips or other useful information to assure maximum function
Caution : Indicates a potential hazard or unsafe practice which, if not avoided,ld cou
result in minor personal injury or product damage.
Warning : Indicates a potential hazard or unsafe practice which, if not avoided, could
result in death or serious injury.
Danger :
Is defined as a source of potential injury to a person.
2.5 Safety Note
2.5.1 Warnings
 Please confirm the colposcope is in normal working condition before operation.
 Keep the camera away from extreme humidity. During operation, avoid liquid splashing.
Otherwise, it may result in failure.
 Never place the camera under a temperature of higher than 60C or direct high
temperature radiation of sensors of thermal therapeutic or microwave devices.
 This is a non-contact type of examination device. Avoid contacting patient during use.
 To ensure safety use, user must provide standard three -prong power socket and ensure
proper connection of earth protection.
 Before turn off power supply of the colposcope, never plug or unplug the video cable.
 Never stare at the illumination lens because it is extremely bright.
2.5.2 Precautions
 Without the written ap proval of the manufacturer, user shall not remove, modify or use
accessories which do not comply with the manufacturer’s requirements.
 When connecting the CABLE, handle it carefully. Align the connectors, do not to plug or
unplug with strong force, otherw ise, the cable and connectors may be damaged, which
will impact the image quality or camera power supply or even the equipment.
9
MedGyn AL-106 Digital Video Colposcope User Manual
 When not in use, the power supply shall be turned off, and the camera shall be covered
to prevent dust.
 To prevent accidental power failure and other potential accidents, use UPS power
system with the colposcope.
 Before cleaning or sterilizing the colposcope, make sure to disconnect the power cable.
Clean and sterilize according to the items specified.
 Do not use the device with other electrical devices.
 Keep cellular phones away from the device.
 When use diskette, CD, or other device to copy data, please be aware of virus.
 Packing materials should be put away from children reach or handled according to
related regulations.
 In case any doubt arises, please consult as soon as possible.
 Please comply with the items stipulated in this manual. This manual is not the
replacement of medical regulations.
 Notes indicate points of particular interest for more efficient and convenient operati on.
 Always take particular notice of the warnings, cautions and notes provided throughout
this manual.
Note :Digital video colposcope is not a treatment device, and it is only a
non- contact examination device.
2.6 Contraindication
No contraindication
10
MedGyn AL-106 Digital Video Colposcope User Manual
3. Camera
This chapter explains the camera composition and installation.
3.1 Composition
AL-106 includes high-resolution color CCD, double circular LED group light source,
control circuit and enclosure. The outer view is as following: Fig.3 -1
①
Fig. 3-1:
①
① — Enclosure
② — LED group light
③ — Switch
④ — Button
③
②
③
④
④
②
⑤ — Handgrip
⑤
⑤
Fig. 3-1 camera composition
3.2 Installation
Thread the camera output cable through the hinge fixed block ② then fix the camera on the
hinge ③ with 2PCS M4 black round head screws.
Fig. 2-2:① — Stand fixing knob
② — Hinge fixed block
③ — Hinge
④ — Fixed block knob
③
③
②
②
④
④
①
①
Fig.3-2 camera installation
11
MedGyn AL-106 Digital Video Colposcope User Manual
3.3 Disassembly
1)
Screw off the 2pcs M4 screws to disassembly the camera.
Fig. 3-3 Camera disassembly
2)
Take off the fixing base with the screw driver;
Fixing base
Lock
Fig.3-4 lock
3)
Then the camera can be took off from the pole.
Note:
Camera case can only be opened by professionals/trained
personnel.
12
MedGyn AL-106 Digital Video Colposcope User Manual
4. Stand
This chapter explains the composition of the stand and installation
4.1 Vertical Stand
4.1.1
Vertical Stand Composition
Vertical stand is used to support the camera, the stand height could be adjusted,
and adjustable range is within 180mm.
①
②
Fig. 3-1: ① —
②—
③—
④—
⑤—
⑥—
4.1.2
Inner pole
Upper stand
Lower stand
Stand base
Fixing nut
Caster
③
④
⑥
Fig. 4-1 Vertical stand
⑤
Assembly
1)
Screw off the nut from the lower pole;
Fig. 4-2
2)
Put the lower pole through the hole of the stand base, and then fasten the stand
and the base with the nut.
Fig 4-3
13
MedGyn AL-106 Digital Video Colposcope User Manual
3)
Put the camera connection cable through the upper pole. (Align gap a with the
screw hole b)
a
b
Fig 4-4
4)
Fix the camera and stand with one M3 flat screw .
M3 screw
Fig 4-5
5)
Get out the spring wire (c) with the hook (d) from the lower pole, and connect it
with camera output cable(e):
e
c
d
Fig 4-6
14
MedGyn AL-106 Digital Video Colposcope User Manual
6)
Loosen the knob (f), pull up the inner pole (g) to the highest position and hold it,
then rotate the upper stand to connect the upper stand with the lower stand.
(Make sure the inner tube does not rotate along with the upper pole):
g
f
Fig 4-7
7)
For the purpose of better cable management, please mount the power box
cover) with M3 screw, the cable will get out through the outlet (h).
h
M3 screw
Fig 4-8
15
MedGyn AL-106 Digital Video Colposcope User Manual
8)
Then the installation is completed. Press the power switch to start work.
Switch
Fig 4-9
16
MedGyn AL-106 Digital Video Colposcope User Manual
4.2 Swing Arm Stand
4.2.1 Swing arm stand composition
17
MedGyn AL-106 Digital Video Colposcope User Manual
4.2.2 Installation steps
1) Unscrew nut from bottom of pole, place pole in hole of rolling base and
tighten nut.
2) Connect horizontal arm to bottom pole. Turn the top sleeve on the vertical
pole counter clockwise to locate screw placement lock. Insert screw labeled
M4x6 in the horizontal arm and tighten. Tighten the sleeve clockwise.
18
MedGyn AL-106 Digital Video Colposcope User Manual
3) Connect swing arm to horizontal arm. Unscrew the cap on the underside of the
horizontal arm and insert a screw to tighten the swing arm in place. Replace cap.
4) Connect camera support arm to swing arm. Unscrew cap on upper side of camera
support arm. Place one step screw in hole and tighten to hold camera support arm
in place. Replace the cap.
19
MedGyn AL-106 Digital Video Colposcope User Manual
5) Put camera cable through hole of support arm. Release knob, put camera
in support, and rotate knob to fasten camera.
6) Place cable into clips as shown in Fig 4-15:
20
MedGyn AL-106 Digital Video Colposcope User Manual
7) Install fixing plate onto swing arm with M4x8 flat head screws, and install
power and signal conversion box onto fixing plate with 4 M3x8 screws.
Fig.4-16 Install power and signal conversion box
8) Connect DB9 end of DB9-MD6 extension cable to interface marked "SIG
NAL", then tighten. Connect MD6 end to camera cable, (carefully align the arrows
before connecting). Place remainder of cable in buckle of swing arm if necessary.
21
MedGyn AL-106 Digital Video Colposcope User Manual
9) If Video signal out is needed, connect composite video cable to interface marked "Video", and
other side to output device. If S-Video signal out is needed, connect S-Video to interface marked
"S-Video" (arrow should face upward) and other end to output device.
10) Unplug power cable; connect one end to power socket of conversion box,
and other side to power socket. Then lock power cable with anti-slip clip.
22
MedGyn AL-106 Digital Video Colposcope User Manual
11) Installation is complete. Press switch to start.
4.3 Precautions
If swing arm needs adjusting, adjust it as follows:
a) When camera is moved to certain positions, but still drops, use special
wrench (an accessory) to rotate weight adjustment nut counter-clockwise.
b) When camera is moved to certain positions, but still goes up, use the special
wrench (an accessory) to rotate weight adjustment nut clockwise.
23
MedGyn AL-106 Digital Video Colposcope User Manual
5. Camera Operation
This chapter explains the function of the camera and its operation
5.1 Button Function
1. Power indicator ①:
When the switch turns on, the indicator ① flash and then becomes green.
2. “Zoom out” ② :
Under auto focus, this button could control the magnification of observed image,
equivalent to local size reduction . Under manual focus, press this button to
decrease the focal distance
3. “Zoom in” ③ :
Under auto focus, this button could control the magnification of observed image,
equivalent to local enlargement. Under manual focus, press this button to
increase the focal distance.
24
MedGyn AL-106 Digital Video Colposcope User Manual
4. “White” ④:
1) Under green filter function, press this button to return to normal white light
display.
2) Press it again, exposure way change to aperture- priority, with screen display
“FA”; press it once again, aperture increase one grade, with screen
display ”FB” press it again to return to auto status.
3) The default value is normal white light, green filter is off.
5. “Green” ⑤:
1) Press this button to activate green filter function, camera will automatically
filtrate the redness and present the morphology of the vascular.
2) There are three grades of filter, press this button once, screen will display
light green filter result; press it ag ain, screen will display normal green filter
image; press it the third time, screen will display dark green filter image.
3) When start up the camera, white light display is set as default, while green
filter function is off.
6. AF/MF ⑥:
1) This button is used to control auto focus and manual focus
2) Press this button, auto focus turns to manual focus. And then, “zoom out”
button is used to manually adjust the far focus. “Zoom in” is used to
manually adjust the near focus, until clear imaged is obtained. Press “focus”
return to auto focus, and then “zoom in” and “zoom out” function returns to
normal.
3) When start up the camera, auto focus is set as default.
7. Display ⑦:
1) Press this button, screen could display magnification information
2) Press it again, the information disappears
3) When start up the camera, no information display is set as default.
8. Time ⑧:
1) Press this button, display “acetic acid reaction timing” mark
2) Press it again, timing mark will disappear
3) When start up the camera, no timing mark is set as default
9. Still ⑨:
Press this button to freeze the image with “Still” mark display on the screen;
press it again, display will return to normal
10. Camera switch ⑩:
Camera switch is used to turn on and turn off the camera power
25
MedGyn AL-106 Digital Video Colposcope User Manual
5.2 Camera Adjustment
Camera direction and height can be adjusted accords to the doctor’s
height during examination. Method is as following:
Fig. 4-2:
① —
② —
③ —
④ —
Stand fixing knob
Hinge fixed block
Hinge
Fixed block knob
③
②
③
②
④
①
④
①
Fig 5-2 Camera Adjustment
1) Counterclockwise loosen the standing fixing knob ① to adjust the
stand to a proper height, then fix the knob ①;
2) Rotate the camera back and forward through the hinge ③;
3) Counterclockwise loosen the fixed block knob ④ to adjust the camera
left-right, then fix the knob.
26
TRUSTED PRODUCTS FOR PREMIER WOMEN’S HEALTHCARE
MEDGYN PRODUCTS, INC.
100 W. Industrial Rd. Addison, IL 60101 USA
t: +1 630.627.4105 toll-free: 800.451.9667
f: +1 630.627.0127