Uploaded by Sophat kean

toshiba-xario-service-manualmaintanance

advertisement
No. 2D730-503E
SERVICE MANUAL
FOR
DIAGNOSTIC ULTRASOUND SYSTEM
!
MODEL SSA-660A
[MAINTENANCE VOLUME]
(2D730-503E)
!
© TOSHIBA MEDICAL SYSTEMS CORPORATION 2004
ALL RIGHTS RESERVED
No. 2D730-503E
IMPORTANT!
1. No part of this manual may be copied or reprinted, in whole or in part,
without written permission.
2. The contents of this manual are subject to change without prior notice
and without our legal obligation.
C-!1
No. 2D730-503E
C-!2
*
No. 2D730-503E
REVISION RECORD
REV.
DATE
REASON
(MM/YY)
INI.
09/'04
/AUTHOR
Mr.
Muto
R-!1
PAGE
SER.
DOC.
CHANGE
D
No.
PRODUC
T.
-------
TMWP7
*
CONTENTS
No. 2D730-503E
Page
- !a -
1.
No. 2D730-503E
INTRODUCTION
1
1.1
Maintenance Menu Configuration (For Users)
1
1.2
Service Menu Configuration (For Service Engineers)
2
2.
OUTLINE OF MENUS
1
2.1
Outline of Points Common to the Maintenance and Service Menus
1
2.2
Outline of the Maintenance Menu
1
2.3
Outline of the Service Menu
2
2.4
Reboot Function
4
3.
MAINTENANCE MENU FUNCTIONS
1
3.1
Activation and Termination
1
3.2
Archiving
2
3.3
System Maintenance
3
3.4
Remote Access Management
8
3.5
Image Quality Report
9
4.
SERVICE MENU FUNCTIONS
1
4.1
Activation
1
4.2
Termination
- !b -
No. 2D730-503E
4
4.3
Information Function
5
4.4
Test Tools Function
83
4.5
Maintenance
93
4.6
Utilities
111
4.7
Shortcuts
103
- !c -
1.
INTRODUCTION
The test system described in this manual is used for maintenance and servicing of diagnostic
ultrasound systems. It consists of a maintenance menu which assists the user in maintenance of
diagnostic ultrasound systems and a service menu that assists the service engineers in servicing
diagnostic ultrasound systems.
Note: The service menu functions are not available to users.
1.1
Maintenance Menu Configuration (For Users)
The maintenance menu is composed of the following 6 items.
⎯
Menu name
Description (use and purpose)
1
Archiving
• Used to regularly back up user-specific data on separate media
as a measure against problems such as HDD failure.
• Allows the use of removable media.
2
System Test
• Used to check if the hardware is operating normally.
• When something appears to be wrong with the operation of a
device, this function can be used to check the status of the
device.
• This function can identify the malfunctioning board.
• The test results can be saved in the HDD according to the
specified settings
3
System Maintenance
• A template is provided for the user's daily check record.
4
Communication
• This function allows the user to transmit requests and
notifications to the service company.
• This function is also used to check and read the response from
the service company.
5
Remote Access
Management
• With this function, the user can permit the service company to
perform remote maintenance on the system via a network or
telephone line.
6
Image Quality Report
• When a problem (such as noise) occurs for an image, the
settings, such as the imaging conditions, can be saved on the
HDD together with the problem image
Note: 1. Data recorded on the HDD must be backed up regularly.
2. Menus 2 and 4 in the table above will be supported in the future.
1-!1
1.2
Service Menu Configuration (For Service Engineers)
The service menu is composed of the following items:
Item
Information
Description (use and purpose)
1) Site
Looking up and updating site data
2) System
Looking up and updating system data
3) HW/SW
a)
Common I/F
Looking up and updating hardware data
b)
Peripherals
Looking up and updating peripheral device
data
c)
Software
Looking up and updating software data
4) Network
a)
Network I/F (TCP/IP)
Looking up and updating network setting data
for the diagnostic ultrasound system
b)
Modem
Looking up and updating modem setting data
c)
IV Setup
Looking up and updating InnerVision setting
data
5) Configuration
Test Tools
Maintenance
a)
DICOM
Basic settings for DICOM communication
b)
Advanced DICOM
Advanced setting for DICOM communication
c)
File Transfer
Basic settings for data transfer
6) Service Config.
Looking up and updating user menu show/hide
settings
1) Diagnostics
Diagnostic ultrasound system testing
2) HDD
Scandisk
3) DICOM
DICOM test
4) Probe*1
Transducer deterioration test
5) Front End*1
Performance check of the front-end analog
sections
1) File Maintenance
a)
User Data Backup/
Restore
b)
Backup and restoration of user-specific data
File Transfer
File transfer using an FTP
a)
Event Viewer
Log display using Event Viewer
b)
Original Log Viewer*1
Log display of service results
2) Log
3) Report
Viewing the information entered using the
Communication user menu and sending
responses.
4) Service History
Looking up and updating service history data
1-!2
Utilities
Short Cut
5) Performance*1
Display of phantom image acquisition
procedure
1) Command Line
Execution of commands
2) Color Adjust*1
Color adjustment
3) Remote Application
Display of pop-up windows on the diagnostic
ultrasound system
1) Smart Backup
Backup of user-specific data and preset data
2) Software Upgrade
Software version upgrade
*1: These functions will be supported in the future.
1-!3
*
2.
2.1
OUTLINE OF MENUS
Outline of Points Common to the Maintenance and Service Menus
(1)
Operation
When the menu screen is opened, a full-screen dialog box for specifying the functions to be
started up is displayed. When service mode is started up, the touch panel screen and the
panel are placed in service mode until service mode is terminated. Usual panel operation
and touch panel screen operation are disabled in service mode.
(2)
Restarting interrupted software
If service mode is started when other software components are in the interrupted status, it
may not be possible to restart the software after exiting service mode. The system may have
to be turned OFF and restarted after servicing is completed.
2.2
Outline of the Maintenance Menu
Operation of the maintenance menu is possible only at the diagnostic ultrasound systems. The
menu is provided as a pop-up menu. The Image Quality Report function, however, is provided in
the OTHER menu on the touch panel.
(1)
Basic maintenance menu window
The maintenance menu window is configured with the following basic format.
Cancel
!
Go
Exit
Each window has [Cancel], [Go], and [Exit] buttons. The buttons have the following
functions:
(a)
[Go]
: Executes the specified process.
(b)
[Cancel] : Stops the process in progress immediately. The process in progress is
returned to the status before execution.
(c)
[Exit]
: Closes the window.
2-!1
(2)
Setting of whether menus are to be displayed
It is possible to set whether or not the System Test, System Maintenance, Image Quality
Report, and Communication menus are displayed. The settings are specified in Service
Config in the service menu (refer to subsection 4.3.8).
In the default status, the System Test and Communication menus are not displayed. These
menus should not be displayed until the corresponding function is supported.
2.3
Outline of the Service Menu
The service menu can be operated in two ways: by displaying it on the system monitor or by
accessing it using a remote PC.
The service menu can be operated in two ways: operation at the system by displaying it on the
system monitor or remote-access operation by displaying it on the InnerVision system.
2-!2
(1)
Service menu window
Service menu main window
Main Menu
Service Software
Information
Site
Test Tools
System
Configration
HW/SW
Network
Service Config
Maintenance
File
Maintenance
HDD
DICOM
Stationing
Application
Remote
Application
Utilities
Log
Report
Service History
Short Cut
Smart
Backup
Diagnostic
Command
Line
Service Exit
Software
Upgrade
Exit
!
Service menu item window: When an item is selected, a window with the following
configuration is displayed.
!
2-!3
2.4
Reboot Function
The system is rebooted automatically in the following cases:
(1)
When system operation becomes unstable
(2)
When the basic system settings are changed
2-!4
*
3.
3.1
MAINTENANCE MENU FUNCTIONS
Activation and Termination
(1)
Activation
Press the MENU switch on the panel, select Maintenance in the pop-up menu, and then
select the desired function in the submenu.
!
(a)
Archiving, System Maintenance, Remote Access Management
These functions are activated by selecting the corresponding function name in the
Maintenance submenu.
(b)
Image Quality Report
This function is activated by pressing [IQ Report] in the OTHER menu on the touch
panel.
(2)
Termination
To terminate a function, select Exit in the dialog box for the function. Refer to the explanation
of termination processing for each function.
3-!1
3.2
Archiving
(1)
Purpose
(a)
Used to regularly back up the user-specific data on separate media as a measure
against problems such as HDD failure.
(b)
The data to be backed up is listed below:
• User-defined measurement data (values defined using Application Preset Editor in Exam
Type Edit in the System Setting pop-up menu)
• User-defined system presets (preset values defined in General Preset, Peripheral System,
OUTPUT, and Exam Information in the System Setting pop-up menu) and Imaging presets
(imaging parameter information that can be saved for each Exam Type displayed by
pressing the Preset switch on the touch panel)
• Service-related presets (preset values defined in the service menu)
Note: Removable media must be used to back up the user-specific data.
The system identification data must be stored on the media. Otherwise, the media is
ejected automatically and the data cannot be saved. Each system requires dedicated
media.
(2)
Display window
!
3-!2
No.
(3)
3.3
Item
Description
1.
Drive
• Used to specify the drive (external media) for making
the backup copy.
2.
Refresh Drive List
• Used to obtain the drive connection data
3.
Initialize Media
• Used to write system identification data onto backup
media. This process cannot be aborted.
• This button is disabled during backup operation
4.
Progress Status Bar
• Indicates the progress of archiving.
5.
[Cancel]
• Aborts backup operation. The data backed up before
the process was aborted cannot be restored (the
directory name is changed).
6.
[Go]
• Starts backup operation.
7.
[Exit]
• Closes the window.
Brief description of the procedure
(a)
Prepare (insert) the media for backup.
(b)
Select a drive.
(c)
Press [Initialize Media].
(d)
Press [Go] in the bottom right part of the window.
(e)
The backup process starts.
(f)
The backup process ends. When the backup data is saved, the media is ejected
automatically from the drive.
(g)
Press [Exit] in the bottom right corner to close the window.
System Maintenance
(1)
Purpose
• Provides a template for the daily check conducted by the user.
• Provides a hard copy of the check record and saves the check record on the HDD.
3-!3
(2)
Display window
!
(a)
Check window
No.
Item
Description
1.
Check items
• Items to be checked.
• In the pull-down menu, select No Execution, Pass,
or Fail. The default is No Execution for all items.
2.
Site and system
data
• Displays the site and system data (data is displayed
automatically).
3.
Date
• Test date (Displayed automatically. The date format
normally used in the system is used.)
4.
Tester Name
• Enter the name of the person doing the test. (If the
person responsible has already been specified, his/
her name appears automatically.)
5.
[Display]
• Displays a test pattern on the monitor.
6.
[Execute]
• Starts the Operation Panel test screen.
7.
[Cancel]
• Initializes the entered data.
8.
[Print]
• Prints out the test results and stores the data on the
hard disk.
9.
[Exit]
• Closes the window.
3-!4
(b)
Monitor CRT test screen
When the [Display] button is pressed, a test pattern is displayed on the monitor for
approximately 5 seconds.
!
3-!5
(c)
Operation Panel test screen
When the [Execute] button is pressed, the following window is displayed. The ROM
and RAM tests are executed automatically, and Button Tests and other tests are
performed interactively.
The highlight moves to the item being executed.
For each test, one of the four statuses (Checking, Pass, Failed, and Indistinct) is
displayed. "Checking" appears while the test is being executed. "Pass" appears upon
normal completion of a test. "Failed" appears upon abnormal termination of a test.
"Indistinct" appears when a test is aborted. ROM and RAM checks cannot be aborted.
!
!
The following window is displayed at the top of the test screen for the switch test.
Clicking this button results in
moving to the previous test.
The status of the previous
test is Indistinct. Note that
this button is invalid in the
case of "Button Test 1".
Clicking this button results in
moving to the next test.
The status of the next test
is Indistinct.
Push the illuminated Button.
<< Back
Next >>
The following window is displayed at the top of the test screen for the keyboard test.
!
Clicking this button
results in returning to
"Button Test 5".
Start keyboard test Type something.
<< Back
3-!6
Next >>
Invalid key
<<Operation Panel test items>>
• ROM
The ROMs in the circuit in the panel are checked automatically.
• RAM
The RAMs in the circuit in the panel are checked automatically.
• Panel switch test
The LEDs on the switches of the panel are lit in sequence and the peripheral circuits for
the switches are checked when the switches corresponding to the lit LEDs are pressed.
The peripheral circuits for the following switches are checked.
a) Button Test1 ⎯⎯⎯⎯⎯⎯⎯ New Patient switch
b) Button Test2 ⎯⎯⎯⎯⎯⎯⎯ TRANSDUCER switch
c) Button Test3 ⎯⎯⎯⎯⎯⎯⎯ MENU switch
d) Button Test4 ⎯⎯⎯⎯⎯⎯⎯ FREEZE switch
e) Button Test5 ⎯⎯⎯⎯⎯⎯⎯ CDI switch
• Keyboard
The items entered from the keyboard are displayed in a dialog box to enable verification of
normal keyboard entry.
(3)
Check items
No.
Check item name
Check item
Check method
1.
Power Supply
Cable
• Area around the power
outlet to which the power
cable is connected
• Visual inspection
2.
Smell/Noise
• Abnormal noise or smell
• Checking for noise and
smell
3.
Air Filter
• Air filter
• Visual inspection
4.
Fixture
• Securing of peripheral
devices
• Visual inspection
5.
Caster
• Caster operation
• Visual inspection
6.
Probe
• Transducer
• Visual inspection (checking
for abnormality in
appearance)
7.
Peripheral
Equipment
• Peripheral devices
• Visual inspection
8.
Image Quality
• Image quality
• Visual inspection (checking
for image abnormalities
before the service menu is
started)
9.
Monitor CRT
• Monitor
• Displaying a test pattern
• Operating panel
• Checking entries made from
the panel
• Displaying the operation
test instruction dialog by
pressing the operation test
button
10. Operation Panel
3-!7
(4)
3.4
Brief description of the procedure
(a)
User check.
(b)
Enter the name of the person who conducted the test.
(c)
Save the test results on an HDD and also make a hard copy.
(d)
Press [Exit] in the bottom right corner to close the window.
Remote Access Management
(1)
Purpose
The user can permit the service company to perform remote maintenance on the system via
a network or telephone line.
(2)
Display window
!
No.
1.
(3)
Item
[Exit]
Description
• Closes the window. Reboots the system.
Brief description of the procedures
Press [Exit] in the bottom right corner of the window to close the window. The system is
rebooted.
3-!8
3.5
Image Quality Report
(1)
Purpose
When an abnormality occurs for an image, this function saves the image and the system
data.
(2)
Display window
No window is provided. The operation status is indicated by the "in-service" display on the
touch panel.
(3)
Brief description of the procedure
No special operation is required.
When a user identifies an abnormal image, the image can be saved in the following manner.
Freeze the image ! Select OTHER menu (touch panel) ! Press [IQ Report] ! The contents
displayed on the TV monitor are saved.
3-!9
*
4.
4.1
SERVICE MENU FUNCTIONS
Activation
(1)
License key for activation
A license key based on a fixed Vendor Host ID is required for activation.
Note: For on-site license key activation, the service engineer prepares the media (CD, MO,
etc.) containing the license key. If such a media is unavailable, the TAC issues a
license key that is valid for one day only.
(2)
Activation processes
<<Sites with a service contract and InnerVision>>
Activation at the diagnostic ultrasound system
(a)
Select [Service] in the pop-up menu to display the corresponding submenu.
(b)
Press "Ctrl + Shift" on the keyboard and press the right trackball button at the same
time.
(c)
The license key entry window appears.
(d)
Enter the license key.
(e)
The service menu is displayed.
Activation from a network PC
(a)
Give the user the following instructions over the telephone.
Select [Service] in the pop-up menu and select Maintenance. Start Remote Access
Management and permit remote access.
(b)
After remote access is permitted, open the license entry window by entering the
address in the browser on the InnerVision system screen.
(c)
The license entry window is displayed on the InnerVision system screen.
(d)
Enter the license key.
(e)
The service menu is displayed on the InnerVision system screen.
<<Sites with a service contract only (no InnerVision)>>
Activation at the diagnostic ultrasound system
•
Operations on the diagnostic ultrasound system are identical with those for sites with a
service contract and InnerVision.
Activation from a network PC
•
Remote access is not supported.
4-!1
Select the desired function
from the menu displayed
by selecting [Service] in
the pop-up menu.
Enter the special
key combination
Activate Remote
Access Management
At the system
Check the service license.
License key entry window
Enter the license key.
Service menu
main window
When it is
acceptable
Maintenance
operations
Service menu
main window
Request the customer
by telephone to permit
remote access.
When it is
acceptable
Check the service license.
Remote access
Enter the license key.
Specify the address in the browser.
License key entry window
!
<<Sites without a service contract>>
Activation act the diagnostic ultrasound system
(a)
Set a media (CD, MO etc.) that is ready for on-site use into the corresponding drive of
the diagnostic ultrasound system.
(b)
Select [Service] in the pop-up menu to display the corresponding submenu.
(c)
Press "Ctrl + Shift" on the keyboard and press the right trackball button at the same
time.
(d)
The license key entry window appears. Press the "License Read" button.
Activation from a network PC
•
Remote access is not possible.
4-!2
(3)
License key entry window:
Enter the license key in this field.
Asterisks are displayed for the
entered characters.
License Read
7CD9CF68768956
Ok
Cancel
Set As Default
!
No.
Items
Description
LM
RM
For sites with a service license (with a service contract and InnerVision or with a service
contract only)
1.
Key entry
• Used to enter the license key (asterisks (*) are
displayed for the lower-order 6 characters entered).
○
○
2.
[License
read]
• Used to read the license key from an external media.
• The media is ejected when the system enters service
mode.
○
X
3.
[License
read]
• The button is not active.
X
○
4.
[Ok]
• Used to proceed to the next step after the key is
entered/read correctly.
• If media is set in the drive, it is ejected.
○
○
5.
[Cancel]
• Used to close the window. The system does not enter
service mode.
• If media is set in the drive, it is ejected.
○
X
6.
[Cancel]
• Used to close the window.
X
○
7.
[Set As
Default
• Check this box to fix the higher-order 14 characters.
○
X
○
-
For sites without a service license (without a service contract)
8.
[Entry
window]
LM: Local mode
RM: Remote mode
• The window is not displayed.
O: Effective
X: Not effective
A license key is a string of 20 alphanumeric characters. The higher-order 14 characters of
the key can be stored in the system by selecting "Set as Default" in the license key entry
window.
Once the higher-order 14 characters are in the system, the service menu can be started by
entering only the lower-order 6 characters.
4-!3
(4)
Service menu window
Main Menu
Service Software
Information
Site
Test Tools
System
Configration
HW/SW
Network
Service Config
Maintenance
File
Maintenance
Log
Report
Short Cut
Command
Line
Stationing
Application
Remote
Application
Exit
Termination
(1)
DICOM
Service Exit
Software
Upgrade
!
4.2
HDD
Utilities
Service History
Smart
Backup
Diagnostic
Normal termination
Select [Exit] in the service menu window.
4-!4
4.3
4.3.1
Information Function
Site
(1)
Purpose
• Looking up and updating site information (information about the location where the
system is installed).
Note: When the HDD is replaced, enter the serial number of the new HDD.
The first character of the serial number must always be replaced with "3".
Example
(2)
:
Serial number shown on the label
Serial number to be entered
A4572001
34572001
Display window
!
4-!5
No.
Item
Description
LM
RM
1.
Site Name
• Used to look up the hospital name.
○
○
2.
Location
• Used to look up the hospital department in which the
system is installed (including installation location).
○
○
3.
Serial Number
• Used to set and look up the system serial number.
○
○
4.
Administrator
• Name
• Used to look up the person responsible for the system at
the hospital.
• Used to look up and update the customer management ID
number.
○
○
• ID
5.
Address
• Used to look up the address data (Hospital, Street, Street
No., Zip Code, Phone No., City, State/Region, and
Country).
○
○
6.
Memo
• Used to enter the hospital data (name of the chief
ultrasonographer etc.).
○
○
7.
[Save]
• Saves the entered data in the system.
○
○
8.
[Cancel]
• Abandons all the data entered in this window and
overwrites it with the data stored in the system.
Note: Canceling after saving data does not restore the data
before saving.
○
○
9.
[Home]
• Returns the system to the service menu main window.
○
○
LM: Local mode
RM: Remote mode
(3)
O: Effective
X: Not effective
Brief description of the procedure
(a)
Enter the customer name and customer ID.
(b)
Enter the hospital data etc. in Memo field.
(c)
Press [Save] to save the entered data.
4-!6
4.3.2
System
(1)
Purpose
Looking up and updating system data.
(2)
Display window
!
4-!7
No.
Item
Description
LM RM
1.
Product Name
System series name (factory preset)
○
○
2.
Product Serial No.
System serial number (factory preset)
○
○
3.
System Use Time
• Displays the time of system use
relative to a specified time period.
○
○
○
○
○
○
• Use
Time
• The use time is displayed in hours, with
the minutes rounded up.
• Set/ • Initializes or sets the use time (unit =
Rese
hour).
t
4.
Probe Use Time
5.
• The system inspection schedule can be
Inspection Schedule
specified.
(Note: This function will be
supported in the
future.)
• Information message
Display
• Date, Target
date
• Hours, Use
Time
• Cumulative ON time for each
transducer.
• Select the transducer from the list box.
The cumulative ON time for the
selected transducer is displayed in
hours (the minutes are rounded up).
• The list box displays only transducers
that have been connected to the
system.
• When an inspection is due, it is
possible to display a message
indicating this when the power is turned
ON. When this check box is checked,
an entry window is displayed
automatically.
• The inspection schedule is specified in
terms of date. Enter the date of the
next inspection.
• The inspection schedule is specified in
terms of use time. Enter the use time
after which the next inspection is to be
performed.
• Every Power
OnTime
• The message is displayed every time
the power is turned ON when an
inspection is overdue.
• Just One Time
• The message is displayed only the first
time the power is turned ON after an
inspection becomes overdue.
6.
Installed from
Date of system installation
○
○
7.
Installed until
Expiration date of lease etc. (This must be
entered even for non-lease systems.)
○
○
8.
Maintenance contract
Presence/absence of maintenance
contract
○
○
9.
InnerVision support
Presence/absence of InnerVision contract
○
○
Saves the entered data in a file.
○
○
10. [Save]
4-!8
11. [Cancel]
Abandons all the data entered in this
window and overwrites it with the data
stored in the system.
Note: Canceling after saving data does
not restore the data before saving.
○
○
12. [Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!9
(3)
(4)
Brief description of the procedure
(a)
Specify the inspection date.
(b)
Enter the lease expiration date.
(c)
Enter whether or not there is a maintenance contract.
(d)
Enter whether there is an InnerVision contract.
(e)
Press [Save] to save the entered data.
Message display function
If an inspection is not performed by the date specified in the Inspection Schedule, a
message is displayed each time the system is started up.
4.3.3
4.3.3.1
HW/SW
Common I/F
(1)
Purpose
Looking up board information. It is possible to save board information that cannot be
detected automatically.
(2)
Display window
Information HW/SW
Information
Site
System
HW/SW
Common IF
Peripherals
Software
Option Install
Network
Configuration
Service Config
Test Tools
Maintenace
Utilities
Short Cut
Home
No
Name
P-No/Revision No (Auto)
P-No/BSM No (Manual)
Report
01
TI
unknown
Automatic
02
TX0
unknown
Automatic
03
TX1
unknown
Automatic
04
RX
unknown
Automatic
05
RC
unknown
Automatic
06
BE
unknown
Automatic
07
VP
unknown
Automatic
08
MA
unknown
Automatic
09
IO
unknown
Automati
10
BP
BP
Manual
11
CB
unknown
Automatic
Cancel
! Ready
4-!10
Save
No
.
Item
LM
R
M
1.
No.
I/F numbers
○
○
2.
Name
Names of hardware items in the system
○
○
3.
P-No./Revision
No. (Auto)
Display of the following board information.
• The P-list number and revision are
displayed for the boards for which
automatic detection is possible (TI, TX0,
TX1, RX,RM, IO, RC, BE, VP, CB, and IP
boards).
○
○
P-No./BSM
No. (Manual)
• The revision and BSM number can be
entered for the boards for which
automatic detection is not possible (BP,
Physio, and Panel boards).
4.
Report
Indicates whether automatic detection is
possible or not. For the PWBs for which
"Manual" is displayed, enter the data
manually.
○
○
5.
[Cancel]
Abandons all the data entered in this window
and overwrites it with the data stored in the
system.
Note: Canceling after saving data does not
restore the data before saving.
○
○
6.
[Save]
Saves the entered data on the HDD. The
automatically detected data is not saved.
○
X
7.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
(3)
Description
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Enter the data manually for PWBs for which "Manual" is displayed.
(b)
Press [Save] to save the entered data.
4-!11
4.3.3.2
Detail
(1)
Purpose
Looking up detailed board information. (This function is applicable only to boards for
which automatic information detection is possible.)
(2)
Display window
Information HW/SW
Information
Site
System
HW/SW
Common IF
Detail
Peripherals
Software
Option Install
Network
Configuration
Service Config
Test Tools
Maintenace
Utilities
Short Cut
!
Common I/F Detail
1 TI
SELECT
1 TI
2 TX0
3 TX1
4 RX
5 RC
6 BE
7 VP
8 MA
9 IO
10 BP
11 CB
12 IP
11 CB
12 IP
13 Physio
14 Panel
Cancel
Ready
Save
No.
Item
Description
LM
R
M
1.
Selection combo
box
Displays the I/F numbers and names for
selection.
Selection is possible only for the boards for
which "Auto" is displayed in the "Report" column
in the Common I/F window.
○
○
2.
Common I/F
Detail
Displays data, such as P-list numbers, that can
be detected automatically.
○
○
3.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
(3)
Home
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Select the desired PWB name from the pull-down menu.
(b)
The detailed information for the selected PWB is displayed.
4-!12
4.3.4
Peripherals
(1)
Purpose
Looking up and updating the peripheral device information.
(2)
Display window
!
No.
Item
LM RM
1.
No.
Number
○
○
2.
Equipment Name/
Memo
Enter the data for the peripheral devices
connected to the system.
○
○
3.
[Cancel]
Abandons all the data entered in this window and
overwrites it with the data stored in the system.
○
○
4.
[Save]
Saves the entered data on the HDD.
○
○
5.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
(3)
Description
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Enter the data for the peripheral devices.
(b)
Press [Save] to save the entered data.
4-!13
4.3.5
Software
(1)
Purpose
Looking up software information
(2)
Display window
Main Menu
Information
Site
System
HW/SW
Common IF
Peripherals
Software
Option Install
Network
Configuration
Service Config
Test Tools
Maintenace
Utilities
Short Cut
Home
Software
Windows 2000
Service Pack 4
Middleware Version : VX02A
Product Software Version : 1.00*R000
Service Software Version : 1.00*R000
Data Version : 1.00*R000
Option Software USHI-770A : Enable
Option Software USDF-770A : Enable
Option Software USFR-770A : Enable
Option Software USDI-770A : Enable
Option Software USDI-772B : Enable
Option Software USSE-770A : Enable
Option Software USAT-770A : Enable
Option Software USIA-770A : Enable
Option Software USPV-770A : Enable
!
Ready
No.
Item
Description
LM RM
1.
Software
Software name and version data. The following
software data is displayed:
• Middleware version
• System software version
• Service software version
• Data version
• Optional software
○
○
2.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!14
4.3.6
4.3.6.1
Network
Network I/F (TCP/IP)
(1)
Purpose
Looking up the network configuration for the diagnostic ultrasound system.
(2)
Display window
!
No.
Item
Description
LM
R
M
1.
Adapter
Network hardware data stored in the system.
The hardware can be selected from the combo
box. When the hardware is selected, all the
hardware data is updated accordingly.
○
○
2.
Obtain an IP
address from
DHCP server
or
Specify an IP
address
Select one of the two options.
If "Obtain..." is selected, "IP address", "Subnet
Mask", and "Default Gateway" cannot be
specified.
○
○
3.
[Cancel]
Abandons all the data entered in this window
and overwrites it with the data stored in the
system.
○
○
4.
[Save]
Saves the settings.
○
X
5.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
Note: When it is necessary to configure the diagnostic ultrasound system for connection
to the hospital network or to the InnerVision system, set the IP address and other
necessary data according to the installation manual.
4-!15
4.3.6.2
Modem (TCP/IP)
(1)
Purpose
Looking up and updating the modem setting information.
(2)
Display window
!
No.
Item
Description
LM
R
M
1.
Modem
Modem name
○
○
2.
Com Port
COM port number
○
○
3.
[Cancel]
All the data entered in this window is
abandoned and replaced with the data stored in
the system.
○
○
4.
Allow remote TCP/
IP clients to access
The access range of the Windows2000 RAS
server can be looked up and updated.
○
○
• this computer
only
• Entire
network
• Limits access to this PC through RAS.
• Expands the range of access to PCs on a
LAN through RAS.
○
○
5.
Allow remote TCP/ • When this is selected, the PC-RAS server
IP clients to request
assigns an address to the client when the
a predetermined
system is connected to the RAS.
address
• When this is not selected, the server allows
connection to the RAS with the address
requested by the client.
• TCP/IP address
pool Begin/End
• Enter the IP addresses to be assigned when
"Allow --" is selected.
6.
[Save]
Saves the entered data ("TCP/IP Address pool"
only) in the system (system reboot required)..
○
○
7.
[Home]
Returns the system to the service menu main
window.
○
○
4-!16
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!17
4.3.6.3
IV-Setup
(1)
Purpose
Looking up and updating the InnerVision setup data. (This data is required for the use
of InnerVision.)
(2)
Display window
!
4-!18
No.
Item
Description
LM
RM
1.
Connecting Device
Method of connecting the system to the TAC center.
Exclusive selection.
○
○
2.
Dialup Router
• IP Address
• Configuration button
Valid only when "Dialup Router" is selected for
Connecting Device.
• IP address of the Config window that is supplied by
the dialup router
• Connects to the dialup router using IE. Thereafter, the
setup program of the dialup router is used.
○
○
3.
Modem (TA)
Valid only when "Modem (TA)" is selected for Connecting
Device.
• Telephone book entry name to be used for making the
call
• Telephone number of the TAC center
○
○
○
○
• Phonebook Entry
• TAC Phone Number
4.
Configuration for
TAC-Center
• TAC IP Address
• Data Receive Dir
• Data Send Dir
• Ftp User Name
• IP address to be used by the TAC center
• Reception directory for the concerned site in the TAC
center
• Transmission directory (for the site) to be used at the
TAC center
• FTP user name to be used for sending data to the
TAC center
• Ftp User Password
• FTP password to be used for sending data to the TAC
center
5.
[Cancel]
Abandons all the data entered in this window and
overwrites it with the data stored in a file.
○
○
6.
[Save]
Saves the entered data in the system.
○
○
7.
[Home]
Returns the system to the service menu main window.
○
○
LM:
RM:
(3)
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Enter the IV-Setup data.
(b)
Press [Save] to save the entered data.
4-!19
4.3.7
Configuration
4.3.7.1
Basic configuration for DICOM communication
Whether DICOM configuration is possible depends on whether the required optional
software, such as USDI-770A or USDI-772B, is installed.
Refer to the table below.
Title
USDI-770A not
installed
USDI-770A installed
USDI-770A and
USDI-772B installed
local AEs
o
o
o
character set
o
o
o
Archive devices
o
o
o
Query/Retrieve
x
o
o
Network nodes
x
Hardcopy devices
x
x
o
hc overview
x
x
o
LUT files
o
o
o
HIS/RIS nodes
x
x
o
MPPS
x
x
o
Advanced DICOM
x
x
o
o
(Setting is possible
for Storage only.)
o
(Setting is possible
for Storage and
Query/Retrieve.)
o: DICOM configuration is possible. x: DICOM configuration is not possible.
• The figure below shows the relationship between the USDI-770A and USDI-772B.
SSA-660A
USDI-770A
Remote Activities
USDI-772B
PRINT SCP
PRINT SCU
Storage SCP
Storage SCU
Storage SCU
Storage SCP
Storage
commitment
SCU
Storage
commitment
SCP
Q/R SCU
Q/R SCP
Storage SCP
Storage SCU
MWM SCU
MWM SCP
!
MPPS SCU
Set through the DICOM Menu
Set through the Advanced DICOM menu
4-!20
MPPS SCP
4-!21
4.3.7.1.1
Local AEs (Application Entity titles)
(1)
Purpose
HIS/RIS and Study Transfer for the diagnostic ultrasound system and the application
entity title for Print are set.
The port number is fixed to 104.
(2)
Display window
!
No.
Item
Description
LM
R
M
1.
HIS/RIS
Setting of the application entity title for the
Modality Worklist Management (MWM) SCU of
the diagnostic ultrasound system.
This item is enabled only when the USDI-772B
is installed.
○
○
2.
Study Transfer
Setting of the application entity title of the Study
Transfer (Storage,Q/R) SCU of the diagnostic
ultrasound system.
○
○
3.
Print
AE title of the Print SCU of the diagnostic
ultrasound system.
○
○
4.
Port
The port number for communication is
displayed.
○
○
5.
Host Name
The computer name set in the main system is
displayed.
○
○
6.
[Save]
Saves the changes in the file (system reboot
required).
○
X
7.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!22
(3)
Help screen
!
(4)
Brief description of the procedure
(a)
Enter each AE title. Note that the titles must be unique within the network.
(b)
Press [Save] to save the data.
4-!23
4.3.7.1.2
Character set
(1)
Purpose
Setting the default character set to be used for DICOM communication and for saving
patient information.
(2)
Display window
!
No.
1.
Item
Description
DICOM default character Default character set
set
• 1-byte characters are used. (Fixed)
• Allow only one
character set
LM
R
M
○
○
• All multiple character
sets
• Multi-byte characters are used.
(Note: This will be supported in the
future.)
2.
Character set
This item is used to specify an additional
character set by selecting the
corresponding character code. (Only 1byte character codes are supported.)
However, this is fixed to ISO_IR 100.
○
○
3.
[Save]
Saves the entered character set (system
reboot required).
○
X
4.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!24
(3)
Help screen
!
(4)
Brief description of the procedure
(a)
Select 1-byte characters or multi-byte characters. (Fixed to "Allows only one
character set".)
(b)
Select the character code. (Fixed to ISO_IR 100)
(c)
Save the settings.
4-!25
4.3.7.1.3
Archive devices
(1)
Purpose
Setting the archive device to be used by the diagnostic ultrasound system. The
standard is magneto-optical disk drive (MOD).
(2)
Display window
!
No.
Check item
Description
LM
R
M
1.
Select logical name
Select the logical name to be created in
this window.
○
○
2.
Logical Name
Enter the logical name to be given to the
archive device in this system.
○
○
3.
Drive Type
Select the type of drive connected to this
system.
○
○
4.
Drive Letter
Select the drive letter corresponding to the
selected drive type.
○
○
5.
[Save]
Saves the settings (system reboot
required).
○
○
6.
[Delete]
Deletes the settings (system reboot
required).
○
X
7.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!26
(3)
Help screen
!
(4)
4.3.7.1.4
4.3.7.1.4.1
Brief description of the procedure
(a)
Enter the logical name to be given to the archive device in this system.
(b)
Select the type of the archive device.
(c)
Select the drive letter for the archive device.
(d)
Save the settings.
Network nodes
Host
(1)
Purpose
Setting the network information related to the DICOM transfer target. For connection
via a network (LAN), set the host name and TCP/IP address. For connection via a
modem (RAS), set the phone number, TCP/IP address, domain name, password,
and login script of the remote host.
Note: Currently, only LAN is supported.
4-!27
(2)
Display window
!
4-!28
No.
Check item
Description
LM
RM
1.
Select Host
Selection of a host that has already been set.
When a name is selected here, the corresponding
host name and IP address that have already been
set are displayed. When these settings are to be
changed, change them and then click the Save
button to save the new settings. To add new
settings, select "define new".
○
○
2.
Host Properties
• Host Name
Setting of the host name
• Searches for the specified computer name (if
applicable) on the network or for the logical
name assigned for convenience (valid only
within the system).
○
○
Select the connection method (LAN or RAS).
To use a LAN (fixed to LAN)
To use a telephone circuit
(Note: This will be supported in the future.)
○
○
(Note: This will be supported in the future.)
○
○
• TCP/IP address
3.
Set Local Host
4.
connected by
When this box is checked, the NetBios computer
name is displayed in the Host Name field.
• LAN
• RAS
5.
Data items required to use
RAS
•
•
•
•
•
•
• Enter the IP address of the transfer target.
Phone No.
PPP Login Script
PPP Domain
PPP Account
PPP Password
Verify PPP Password
•
•
•
•
•
•
Phone number of the transfer target
Login script, if applicable
Domain name of the transfer target
Login name (account) of the transfer target
Password to connect to the transfer target
Password to verify connection to the transfer
target
6.
[Save]
Saves the entered data(system reboot required).
○
X
7.
[Delete]
Deletes the settings(system reboot required).
○
X
8.
[Find]
Searches for the specified host name on the
network.
○
○
9.
[Test]
Searches for the specified TCP/IP address on the
network.
Executes the Ping test.
○
○
10.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!29
(3)
Help screen
!
(4)
Brief description of the procedure
(a)
Enter the logical name (computer name) of the DICOM transfer target.
(b)
Select the connection method.
(c)
For LAN connection, enter the IP address only. For RAS connection, enter the
data items required to use RAS. (Only LAN connection is supported.)
(d)
Save the entered data.
(e)
Press [Test] to confirm the DICOM transfer target exists on the network.
4-!30
4.3.7.1.4.2
Remote
(1)
Purpose
Setting the AE data (title and port number) and DICOM service information of the
DICOM transfer target. The image transfer syntax can be selected.
If the USDI-770A only is installed, the DICOM transfer target is remote. If the
USDI-772B is also installed, the transfer target for Storage is Advanced DICOM and
the transfer target for Query/Retrieve is remote.
(2)
Display window
Main Menu
!
Information
Site
System
HW/SW
Network
Configuration
DICOM
local AEs
character set
archive devices
network nodes
hosts
remote
Query and Retrieve
hardcopy devices
hc overview
LUT files
HIS/RIS nodes
MPPS
Advanced DICOM
File Transfer
Service Config
Test Tools
Maintenace
Home
Select Logical Name General
logi
Logical Name
logi
Host Selection
qqq
Application Entity Properties
Select AE Title
define new
aeae
AE Title
aeae
Port Number
1234
Verification
Supported Entity Properties
Image Tranfer (Store)
Byte Order
Implicit Little Endian
Explicit Little Endian
Structure Report Properties
Standard
Ready
Private
Save
4-!31
Delete
No.
Item
Description
LM
RM
1.
Select Logical Name
Selection of a logical name that has already been
set.
When a name is selected here, the information for
the corresponding remote system that has already
been set is displayed. When these settings are to
be changed, change them and then click the Save
button to save the new settings. To add new
settings, select "define new".
○
○
2.
Logical Name
Setting for the logical name of the remote system
(The name set here is displayed on the system.)
○
○
3.
Host Selection
Selection of the host of the remote system (The
host that is set in the DICOM-network nodes-host
page should be selected.)
For Storage, be sure to select the host name of this
system.
For Q/R, select the host name of the
communication partner.
○
○
4.
Default Node
Selection of the default node (Do not check this
box.)
○
○
5.
Select AE Title
Selection of an AE title that has already been set.
When an AE title is selected here, the AE title and
port No. that have already been set for and
correspond to the logical name are displayed.
When these settings are to be changed, change
them and then click the Save button to save the
new settings. If no settings have been entered yet,
only "define new" is displayed.
○
○
6.
AE Title
Setting of the AE title.
○
○
7.
Port Number
Setting of the port number.
For Storage, the port number is fixed to 2000.
For Q/R, set the port number of the communication
partner.
○
○
8.
Add/Del
Button for adding/deleting the AE title, port No.,
and Image Transfer information that correspond to
the logical name (Do not use this button.)
○
X
9.
Verification
Button for executing verification for the remote
system.
○
X
10.
Image Transfer (Store)
Setting of the service class support for Image
Transfer (Store)
○
○
11.
Byte Order/Compression
Setting for the Transfer Syntax
(Only Implicit VR Little Endian and Explicit VR Little
Endian of Byte Order are supported. Explicit Big
Endian is not supported and must not be set.)
If data compression is necessary, set it in
Advanced DICOM. (Do not select Compression on
this page.)
○
○
12.
Structure
Properties
Selection for transfer of the Structure Report data
Standard: Check this box to transfer the data using
the SR standard protocol (default).
Private: Check this box to transfer the data using
the Toshiba protocol.
○
○
Report
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!32
No.
Item
Description
LM
RM
13.
Save
Button for saving the settings (system reboot
required)
○
X
14.
Delete
Button for deleting the settings (system reboot
required)
○
X
15.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
(3)
Local mode
Remote mode
O: Effective
X: Not effective
Help screen
!
4-!33
4.3.7.1.5
4.2.7.1.5.1
Query/Retrieve
Host
(1)
Purpose
Setting the DICOM Query/Retrieve host.
(2)
Display window
!
No.
Item
Description
LM
R
M
The list of the host names stored using
this window is displayed in the pull-down
menu.
○
○
○
○
1.
Select Host
2.
Host Properties
• Host
Name
• TCP/
IPaddress
3.
Set Local Host
When this box is checked, the NetBios
computer name is displayed in the Host
Name field.
○
○
4.
connected Host
Select the connection method (LAN or
RAS).
○
○
• Enter the logical name of the
communication partner on the
network (computer name) or the
logical name assigned for
convenience (effective only in this
system).
• Enter the IP address of the
communication partner on the
network.
• To use a LAN (fixed to LAN at
present)
• To use a telephone line (Note: This
will be supported in the future.).
4-!34
No.
5.
Item
Data items required to use
RAS
Description
(Note: This will be supported in the future.)
• Phone NO.
• Phone number of the communication partner
• PPP Login Script
• Login script, if applicable
• PPP Domain
• Domain name of the communication partner
• PPP Account
• Login name (account) of the communication
partner
• PPP Password
• Password to connect to the communication
partner
• Verify PPP Password
• Password to verify connection to the
communication partner
LM
RM
○
○
6.
[Save]
Saves the settings (system reboot required).
○
X
7.
[Delete]
Deletes the settings (system reboot required).
○
X
8.
[Find]
Searches for the specified computer name
(if applicable) on the network.
○
○
9.
[Test]
Searches for the specified TCP/IP address on the
network.
○
○
10.
[Home]
Returns the system to the service menu main
window.
○
○
4-!35
4.3.7.1.5.2
Remote
(1)
Purpose
Setting the DICOM Query/Retrieve partner and setting the detailed data required to
communicate with the host specified in subsection 4.3.7.1.5.1.
(2)
Display window
!
No.
Item
1.
Select Logical Name
2.
General
3.
4.
LM
R
M
The list of the logical names of the
communication partners stored using this
window is displayed in the pull-down
menu.
○
○
○
○
○
○
○
○
• Logical
• Enter the logical name given for
convenience (effective only in this
system).
• Host
Selection
• Select the host name specified in
subsection 4.3.7.1.5.1
Application
Supported
Description
Entity
• Properties
• Select AE
Title
• The list of the AE titles of the
communication partners stored using
this window is displayed in the pulldown menu.
• AE Title
• Enter the AE title of the
communication partner.
• Port
Number
• Enter the port number used by the AE
title.
DICOM
• services
• Check this box to perform image
transfer.
• Image
Transfer
(store)
• Byte
Order
• Select the DICOM data format to be
used in image transfer.
4-!36
5.
[Save]
Saves the settings (system reboot
required).
○
X
6.
[Delete]
Deletes the settings (system reboot
required).
○
X
7.
[Home]
Returns the system to the service menu
main window.
○
○
4-!37
4.3.7.1.6
4.3.7.1.6.1
Hardcopy devices
Host
(1)
Purpose
Setting the network information related to hardcopy devices (imager and/or printer).
For connection via a network (LAN), set the host name and TCP/IP address. For
connection via a modem (RAS), set the phone number, TCP/IP address, domain
name, password, and login script of the remote host.
Note: Currently, only LAN is supported.
(2)
Display window
!
4-!38
No.
Item
1.
Select Host
2.
Host Properties
• Host
Name
Description
L
M
R
M
Selection of a host that has already
been set. When a name is selected
here, the corresponding host name and
IP address that have already been set
are displayed. When these settings are
to be changed, change them and then
click the Save button to save the new
settings. To add new settings, select
"define new".
○
○
○
○
• Enter the logical name (computer
name) of the hardcopy device
selected in "Select Host" field or the
logical name assigned for
convenience (valid only within the
system).
Up to 14 1-byte characters can be
entered.
• TCP/IP
Address
• Enter the IP address of the
hardcopy device selected in "Select
Host" field.
3.
Set Local Host
When this box is checked, the NetBios
computer name is displayed in the Host
Name field.
○
○
4.
connected by
Select the connection method (LAN or
RAS).
○
○
○
○
•
LA
N
•
RA
S
5.
Data items required to
use RAS
• To use a LAN (fixed to LAN)
• To use a telephone circuit
(Note: This will be supported in the
future.)
(Note: This will be supported in the
future.)
• Phone No.
• Phone number of the transfer target
• PPP Login
Script
• PPP Domain
• Login script, if applicable
• Domain name of the transfer target
• PPP Account
• PPP
Password
• Verify PPP
Password
• Login name (account) of the transfer
target
• Password to connect to the transfer
target
• Password to verify connection to the
transfer target
6.
[Save]
Saves the entered data(system reboot
required).
○
X
7.
[Delete]
Deletes the settings(system reboot
required).
○
X
8.
[Find]
Searches for the specified computer
name (if applicable) on the network.
○
○
4-!39
9.
[Test]
10. [Home]
LM:
RM:
Local mode
Remote mode
Checks whether the system can
connect to the address set for "TCP/IP
address" on the network. Executes the
Ping test.
When "RAS" is selected for "connected
by", nothing is performed.
○
○
Returns the system to the service menu
main window.
○
○
O: Effective
X: Not effective
4-!40
(3)
Help screen
!
4-!41
4.3.7.1.6.2
Remote
(1)
Purpose
Setting the AE data (title and port number) of the hardcopy device (imager and/or
printer).
(2)
Display window
!
No.
Check item
Description
1.
Select Logical Name
2.
General
LM:
RM:
Selection of a logical name that has
already been set. When a name is
selected here, the information for the
corresponding remote system that has
already been set is displayed. When
these settings are to be changed,
change them and then click the Save
button to save the new settings. To add
new settings, select "define new".
• Logical
Name
• Setting of the logical name of the
remote system. The same name must
be used in Logical Name, AE Title,
and Hardcopy Device.
Up to 20 1-byte characters can be
entered.
• Host
Selection
• Select the printer host name set in
"Host" (in "Hardcopy devices").
Local mode
Remote mode
O: Effective
X: Not effective
4-!42
LM
R
M
○
○
○
○
No.
3.
Item
Description
Application Entity Properties
Select AE Title
Selection of an AE title that has already been set.
When an AE title is selected here, the AE title and
port No. that have already been set for and
correspond to the logical name are displayed.
When these settings are to be changed, change
them and then click the Save button to save the
new settings. If no settings have been entered yet,
only "define new" is displayed.
• AE Title
• Enter the AE title of the transfer target
Up to 16 1-byte characters can be entered
• Port Number
• Enter the port number used by the AE title.
LM
RM
○
○
4.
[Verification]
Executes verification of the specified host.
○
○
5.
[Add]/[Del]
Button for adding/deleting the AE title and port No.
that correspond to the logical name
○
X
6.
[Save]
Stores the data (system reboot required).
○
X
7.
[Delete]
Deletes the stored logical name (system reboot
required).
○
X
8.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
(3)
Local mode
Remote mode
O: Effective
X: Not effective
Help screen
!
4-!43
4.3.7.1.6.3
Details
(1)
Purpose
Setting individual parameters supported by the hardcopy devices (imager and/or
printer), such as the film type and film destination.
(2)
Display window
!
4-!44
No.
Item
Description
LM
R
M
1.
Select HC device
Selection of the HC drive name that has
already been set. When a name is
selected here, the information for the
corresponding remote system that has
already been set is displayed. When
these settings are to be changed,
change them and then click the Save
button to save the new settings. To add
new settings, select "define new".
○
○
2.
HC Device
Setting of the HC device name of the
remote system (The name set here is
displayed on the system.) The same
name must be used in Logical Name, AE
Title, and Hard copy Device.
Check the box to set the entered
hardcopy device as the default. (This
function is not currently supported.)
○
○
○
○
○
○
Default Camera
3.
General Settings
• Type
• Select the type of printer.
(Normally, select DICOM Printer.)
• Class
• Specify the class of the cameraspecific filming properties and the
hardcopy device. (Refer to table 1 for
the supported hardcopy devices.)
• DICO
M
Node
• Enter the logical name set in
"Remote" (in "Hardcopy devices").
• Locatio • Enter a comment describing the
n
location.
• Comm
ent
4.
5.
• Enter a comment if necessary.
Filming Properties
• Hold printed
film jobs
• Select the number of jobs that can be
held.
• Pixel Size
[1/1000 mm]
• Enter the pixel size.
• Film Sheet
Formats
• Select the film sheet format.
• Film Sheet
Format
• Define a new film sheet format.
• Number of
Pixel
• Define the number of pixels
(resolution).
• Medium type
• Select the film type to be used.
• Film
destination
• Select the film supply method.
• Transformation
• Select the interpolation method to be
used for output to the film
[Add]/[Del]
Adds or deletes a film sheet format.
4-!45
6.
[Save]
Saves the settings (system reboot
required).
○
X
7.
[Delete]
Deletes the settings (system reboot
required).
○
X
8.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!46
Table 1 List of supported hardcopy devices
•
•
•
•
•
•
•
•
)
•
•
•
•
•
•
•
•
•
(3)
agfadrystar2000Color
agfadrystar2000
agfadrystar3000
agfalr2100
agfalr3300
agfalr5200
codonicsnp1660(1600)
codonicsnp1660Mono(1600
FujiDRYPIX1000
Fujifl_dp_3543
FujiFNPS551
imation8100
imation8300
imation8500
imation8700
imation959
imation969HQHQS
Help screen
!
4-!47
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Kodak1120
Kodak190
Kodak2180
KonicaDRYPRO722
KonicaLi10A
KonicaLi1417DH
KonicaLi21
KonicaLi62P
KonicaLi7Li7ALi7DDLi8
Seiko1720DColor
Seiko1720DMono
SonyUPAD3Color
SonyUPAD3Mono
Sterling1417
Steringinkjet400
Steringlp300400
4.3.7.1.7
HC overview
(1)
Purpose
Setting the default hardcopy device (imager and/or printer).
(2)
Display window
!
No.
Item
Description
LM
R
M
1.
Select logical name
Select the HC device name from among
the HC device names.
○
○
2.
Name
The HC device name set in "Details" (in
"Hardcopy devices") is displayed.
○
○
3.
Type
The printer type set in "Details" (in
"Hardcopy devices") is displayed.
○
○
4.
Comment
The comment entered in "Details" (in
"Hardcopy devices") is displayed.
○
○
5.
Location
The location entered in "Details" (in
"Hardcopy devices") is displayed.
○
○
6.
Hold printed film jobs
The number of film jobs set in "Details" (in
"Hardcopy devices") is displayed.
○
○
7.
Default Camera
If this box is checked, it indicates that the
printer is set as default in "Details" (in
"Hardcopy devices"). (This will be
supported in the future.)
○
○
8.
Network access [Test]
Checks whether the corresponding
hardcopy device is connected on network.
Executes the Ping test.
○
○
9.
DICOM access [Test]
Checks the AE access.
Executes the DICOM Validation test.
○
○
10.
[Save]
Saves the settings (system reboot
required).
○
X
11.
[Home]
Returns the system to the service menu
main window.
○
○
4-!48
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!49
(3)
Help screen
!
4-!50
4.3.7.1.8
4.3.7.1.8.1
LUT files
Import
(1)
Purpose
Adding or removing a look-up table (LUT).
(2)
Display window
!
No.
1.
Check item
Description
Import LUT
Source
Used to import an external LUT.
• drive
• Drive in which the desired LUT data is
stored
• path
• Directory in which the desired LUT
data is stored
Destination
• LUT
directory
LM
R
M
○
○
○
X
○
X
○
○
• Import destination of the external LUT
data
[copy] Executes import.
2.
Remove LUT
• LUT
directory
• System LUT data directory to be
removed
• LUT file
• LUT data file name
[remove] Executes removal.
3.
LM:
RM:
[Home]
Local mode
Remote mode
Returns the system to the service menu
main window.
O: Effective
X: Not effective
4-!51
(3)
Help screen
!
4-!52
4.3.7.1.8.2
Set
(1)
Purpose
Selecting a LUT and setting the maximum density.
(2)
Display window
!
No.
Check item
Description
LM
R
M
1.
Image type
Specify the LUT data type and density
for each image.
○
○
2.
[Save]
Saves the settings (system reboot
required).
○
X
3.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!53
(3)
Help screen
!
4-!54
4.3.7.2
HIS/RIS nodes
This function is enabled only when the USDI-772B is installed.
4.3.7.2.1
Host
(1)
Purpose
Setting the network information of the host (the station from which the examination
information is obtained).
For network connection (LAN), set the host name and TCP/IP address.
Modem connection is not supported.
(2)
Display window
!
4-!55
No.
Item
1.
Select Host
2.
Host Properties
• Host Name
• TCP/IP
address
3.
Set Local Host
4.
Connected by
LM
R
M
Selection of a host that has already been
set. When a name is selected here, the
corresponding host name and IP address
that have already been set are displayed.
When these settings are to be changed,
change them and then click the Save
button to save the new settings. To add
new settings, select "define new".
○
○
○
○
○
○
○
○
○
○
Enter the logical name (computer name)
of the transfer target, or the logical name
given for convenience (valid only within
the system).
Enter the IP address of the transfer target.
When this box is checked, the NetBios
computer name is displayed in the Host
Name field.
•
•
5.
Description
LA
N
To use a LAN (fixed to LAN)
To use a telephone circuit
(Note: This will be supported in the
future.)
RA
S
Data items required to
use RAS
(Note: This will be supported in the
future.)
• Phone No.
Phone number of the transfer target
• PPP Login
Script
Login script, if applicable
• PPP Domain
Domain name of the transfer target
• PPP Account
Login name for connection to the transfer
target
• PPP Password
Password for connection to the transfer
target
• Verify PPP
Password
Password to verify connection to the
transfer target
6.
[Save]
Saves the settings (system reboot
required).
○
X
7.
[Delete]
Deletes the settings.
○
X
8.
[Find]
Searches for the specified computer name
(if applicable) on the network.
○
○
9.
[Test]
Searches for the specified TCP/IP
address on the network.
Executes the Ping test.
○
○
10.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!56
4-!57
(3)
Help screen
!
(4)
Brief description of the procedure
(a)
Enter the logical name (computer name) of the station with which the system
communicates through the network.
(b)
Select the connection method.
(c)
For LAN connections, set the IP address only. For RAS connections, set the
items required to use RAS. (Currently, only LAN connection is supported.)
(d)
Save the settings.
(e)
Press [Test] to confirm that the communication partner exists on the network.
4-!58
4.3.7.2.2
Remote
(1)
Purpose
Setting the DICOM transfer target and setting the detailed data required to
communicate with the HIS/RIS station specified in "Host" in the previous subsection.
(2)
Display window
!
4-!59
No.
Item
1.
Select Logical Name
2.
General
3.
4.
Description
LM
R
M
Selection of a logical name that has
already been set. When a name is
selected here, the information for the
corresponding remote system that has
already been set is displayed. When
these settings are to be changed, change
them and then click the Save button to
save the new settings. To add new
settings, select "define new".
○
○
○
○
○
○
Setting of the MWM method (active MWM
or passive MWM)
Check the box for Basic Worklist only.
○
○
• Logical
Name
Enter the logical name.
• Host
Selection
Select the HIS/RIS station specified in
"Host" in the previous subsection.
Application Entity
Properties
• Select AE
Title
Selection of an AE title that has already
been set. When an AE title is selected
here, the AE title and port No. that have
already been set for and correspond to the
logical name are displayed. When these
settings are to be changed, change them
and then click the Save button to save the
new settings. If no settings have been
entered yet, only "define new" is
displayed.
• AE Title
Enter the AE title of the transfer target.
• Port
Number
Enter the port number used by the AE title.
Supported DICOM
services
• Basic
Worklist
5.
[Add]/[Del]
Adds or deletes the AE title, port number,
and Supported DICOM services
corresponding to the selected logical
name.
○
X
6.
[Verification]
Executes DICOM Verification (C-ECHO).
○
○
7.
[Save]
Saves the settings. This button becomes
effective when any of above items is set
(system reboot required).
○
X
8.
[Delete]
Deletes the stored data. This button
becomes effective when Logical Name is
selected (system reboot required).
○
X
9.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!60
(3)
Help screen
!
4-!61
4.3.7.2.3
HIS/RIS
(1)
Purpose
Selecting the AE title of the host (the station from which the examination information is
obtained), and setting the timing when the information is obtained from the Basic
Worklist SCP.
(2)
Display window
!
4-!62
No.
1.
Item
Description
Application Entity Title
Selection
• Basic Worklist
SCP
2.
Select a predefined worklist.
LM
R
M
○
○
○
○
Select the AE title (that supports Basic
Worklist) to be actually connected from
among the AE titles set for Basic Worklist
in the HIS/RIS nodes/remote page. If no
AE title is set in the HIS/RIS nodes/remote
page, the message "no Application Entity
selected" is displayed.
Basic Worklist SCP
Query Properties
• Query Waiting Time
Enter the wait time between the query for
the scheduled examinations and the query
response.
• Max Query Match
Number
Enter the maximum number of scheduled
examinations to be obtained (maximum
200).
• Query Interval
Enter the polling interval for automatic
periodic execution of MWM. (Note: This
function will be supported in the future.)
3.
[Save]
Saves the settings (system reboot
required).
○
X
4.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!63
4.3.7.2.4
MPPS
4.2.7.2.4.1 Host
(1)
Purpose
Setting the information of the MPPS station.
(2)
Display window
!
No.
Item
Description
1.
Select Host
The list of the host names stored using
this window is displayed in the pull-down
menu.
2.
Host Properties
• Host Name
• TCP/IP
address
3.
Set Local Host
4.
Connected by
Enter the logical name of the
communication partner on the network
(computer name) or the logical name
assigned for convenience (effective only in
this system)..
LM
R
M
○
○
○
○
○
○
○
○
Enter the IP address of the
communication partner on the network.
When this box is checked, the NetBios
computer name is displayed in the Host
Name field.
•
•
LA
N
To use a LAN.
To use a telephone line (Note: This will be
supported in the future.)
RA
S
4-!64
No.
5.
Item
Data items required to use
RAS
Description
(Note: This will be supported in the future.)
• Phone No.
Phone number of the communication partner
• PPP Login Script
Login script, if applicable
• PPP Domain
Domain name of the communication partner
• PPP Account
Login name (account) of the communication
partner
• PPP Password
Password to connect to the communication partner
• Verify PPP Password
LM
RM
○
○
Password to verify connection to the
communication partner
6.
[Save]
The list of the host names stored using this window
is displayed in the pull-down menu.
○
○
7.
[Delete]
Saves the settings (system reboot required).
○
X
8.
[Find]
Deletes the settings (system reboot required).
○
X
9.
[Test]
Searches for the specified TCP/IP address on the
network.
○
X
10.
[Home]
Returns the system to the service menu main
window.
○
X
4-!65
4.3.7.2.4.2 Remote
(1)
Purpose
Setting the DICOM communication partner and setting the detailed data required to
communicate with the MPPS specified in subsection 4.3.7.2.4.1.
(2)
Display window
!
No.
Item
1.
General
• Logical Name
Description
Enter the logical name given for
convenience (effective only in this
system).
LM
R
M
○
○
○
○
• Host Selection Enter the HIS/RIS name specified in
subsection 4.3.7.2.4.1.
2.
Application
Entity
The list of the AE titles of the
Properties
• Select AE Title communication partners stored using this
window is displayed in the pull-down
menu.
• AE Title
Enter the AE title of the communication
partner.
• Port Number
Enter the port number used by the AE title.
3.
[Save]
Saves the settings (system reboot
required). This is enabled when at least
one data is set.
○
X
4.
[Delete]
Deletes the settings (system reboot
required). This is enabled when the data
for Logical Name is set.
○
X
5.
[Home]
Returns the system to the service menu
main window.
○
○
4-!66
4.3.7.3
Advanced DICOM
4.3.7.3.1
Main Configuration
(1)
Purpose
Setting the AE titles, port numbers, etc. of Advanced DICOM and Basic DICOM in order
to connect them to each other.
(2)
Display window
!
4-!67
No.
Item
Description
LM
R
M
1.
AE Title of Advanced
DICOM
Enter the AE title of Advanced DICOM.
Set the AE title specified in the Study
Transfer field in the window displayed by
selecting DICOM ! local AEs (subsection
4.3.7.1.1).
○
○
2.
Advanced DICOM SCP
Port Number
Port number of Advanced DICOM.
This is fixed to 2000.
○
○
3.
Basic DICOM Host
Name
Enter the Xario host name.
○
○
4.
Basic DICOM Storage
SCP Port Number
TCP/IP port number listened to by
Advanced DICOM in the Local Protocol
function used between Advanced DICOM
and Basic DICOM.
This is fixed to 3000..
○
○
5.
Basic DICOM Storage
SCO Port Number
TCP/IP port number listened to by Basic
DICOM in the Local Protocol function
used between Advanced DICOM and
Basic DICOM.
This is fixed to 5000.
○
○
6.
Remaining Free Space
on Hard Disk (M)
Remaining available space. This is fixed
to 1000 MB.
○
○
7.
CD-R Driver
Drive name for CD-R. This is fixed to G.
○
○
8.
MO Driver
Drive name for MO. This is fixed to F.
○
○
9.
Cache 1
Drive name for cache for CD writing. This
is fixed to U.
○
○
10.
Cache 2
Drive name for the USB HD.
○
○
11.
Enable Removable
Media
Be sure to check this box.
○
○
12.
[Save]
Saves the settings.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!68
4.3.7.3.2
DICOM Destination
(1)
Purpose
For setting Basic DICOM Storage SCP and Remote (Storage, Storage Commitment,
MPPS).
(2)
Display window
!
4-!69
No.
Item
Description
LM
RM
1.
Number of Configured
Destination
The total number of destinations that are currently
set is displayed.
○
○
2.
Select AE Title
Select the AE title of the destination.
○
○
3.
AE Title
Enter the AE title of the destination.
Up to 16 characters can be entered using
alphanumeric characters and underscores (_).
○
○
4.
Host
Enter the host name of the destination.
Set the host name specified in the window
displayed by selecting DICOM ! network nodes !
host (subsection 4.3.7.1.4.1).
○
○
5.
Port
Enter the port number of the destination.
○
○
6.
Alias
Enter an alias (an alternative name for the
destination or a comment that identifies the
destination).
○
○
7.
Packet Size
Enter the packet size.
○
○
8.
[First]
Jumps to the top of the list.
○
○
9.
[Prev]
Goes to the previous list.
○
○
10.
[Next]
Goes to the next list.
○
○
11.
[Last]
Jumps to the bottom of the list.
○
○
12.
[Add]
Adds a new destination.
○
○
13.
[Delete]
Deletes a destination.
○
○
14.
[Save]
Saves the settings.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!70
4.3.7.3.3
Archives
(1)
Purpose
For setting Basic DICOM Storage SCP and Remote Storage SCP.
(2)
Display window
!
4-!71
No.
Item
Description
LM
RM
1.
Number of Archives
The total number of currently set Storage
destinations is displayed.
○
○
2.
Number of Retries
Set the number of retries for Storage and MPPS
processing.
○
○
3.
Commitment association
timeout (sec)
Set the Storage Commitment timeout value.
○
○
4.
Archive Number
Set the list number of the Storage partner.
○
○
5.
Called AE Title
Select the AE title of the destination from the AE
title list set on the DICOM Destinations page.
○
○
6.
Description
Enter a comment.
○
○
7.
AE Title
Enter the AE title of the destination.
Set the AE title of the Storage destination specified
in the window displayed by selecting DICOM !
network nodes ! remote (subsection 4.3.7.1.4.2).
○
○
8.
Compression
Select the compression method from among:
○
○
•
•
•
•
•
•
As it
YBR_FULL
YBR_FULL + RLE
YBR_FULL_422 + JPEG
RLE
YBR_PARTIAL_422 + JPEG
9.
JPEG Compression Ratio
Set the compression ratio for JPEG compression.
This field is enabled only when JPEG compression
is selected for "Compression Type".
○
○
10.
Timeout (sec)
Enter the timeout value for the association
response from the destination.
○
○
11.
Storage Commitment
Check this box when Storage Commitment is
executed automatically after Storage processing.
○
○
12.
USS = >Retired US
Check this box when US Image is converted to
Retired US Image.
○
○
13.
Additional Script
Select the data conversion script. Normally, select
"Blank.scr".
○
○
14.
[Prev]
Goes to the previous list.
○
○
15.
[Next]
Goes to the next list.
○
○
16.
[Add]
Adds a new destination.
○
○
17.
[Delete]
Deletes a destination.
○
○
18.
[Save]
Saves the settings.
○
○
19.
[Test Association]
Executes verification (C-ECHO).
○
○
This should be performed for the set destination.
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!72
4.3.7.3.4
MPPS
(1)
Purpose
Setting the AE title of the destination to which the examination start and end
notifications are sent.
(2)
Display window
!
No.
Item
Description
LM
R
M
1.
Enable Modality
Performed Procedure
Step (MPPS)
Selection of whether MPPS is set to
effective or not effective.
Check this box to make MPPS effective.
○
○
2.
AE Title of MPPS
Selection of the MPPS connection
destination
Select one from the AE title list set in the
DICOM Destination page.
○
○
3.
[Save]
Saves the settings.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!73
4.3.7.3.5
Logging
(1)
Purpose
Setting the method for acquiring the DICOM-related logs.
(2)
Display window
!
No.
Item
Description
LM
R
M
1.
DICOM Logging
Log setting for Merge.log
Checked
: All operations are logged.
Not checked : Only the operations in
which an error occurs are
logged.
○
○
2.
General Logging
Log setting for other logs
Checked
: All operations are logged.
Not checked : Only the operations in
which an error occurs are
logged.
○
○
3.
Maximum Size [KB]
Specify the maximum size of the log file.
○
○
4.
Save
Saves the settings.
○
○
5.
Download Log Files
The downloaded log files are listed in this
area.
When each hyperlink is clicked, the
corresponding log window is opened and
the contents of the file are displayed.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!74
4.3.7.3.6
Import/Export
(1)
Purpose
This procedure should not be performed.
(2)
Display window
!
No.
Item
Description
LM
RM
1.
IMPORT TO APLIO
Enter the file name to be imported (include
the extension).
○
○
2.
Browse...
Used to display a list of file names.
○
○
3.
File name at Aplio
Enter the file name to be saved in Xario
(include the extension).
○
○
4.
Copy to Aplio
Used to import the file specified in item 3
above.
○
○
5.
EXPORT FROM APLIO
The files registered in Xario are listed in this
area.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!75
4.3.7.4
File Transfer
Basic setup for offline data transfer
4.3.7.4.1
Targets
(1)
Purpose
Setting the network data of a host that supports data transfer to the system. (This
setting is required to transmit/receive files via a LAN off line.)
(2)
Display window
!
4-!76
No.
Check item
Description
1.
Select Host
The list of the hosts stored using this window is
displayed in the pull-down menu.
2.
Host Properties
• Host Name
• Enter the logical name (computer name) of the
transfer target, or the logical name given for
convenience (effective only in this system).
• TCP/IP address
3.
Set Local Host
4.
connected by
○
○
○
○
○
○
○
○
• Enter the IP address of the transfer target.
Select the connection method (LAN or RAS).
LAN • To use a LAN
RAS • To use a telephone circuit
(Note: This will be supported in the future.)
Data items required to use
RAS
•
•
•
•
•
•
RM
When this box is checked, the NetBios computer
name is displayed in the Host Name field.
•
•
5.
LM
Phone No.
PPP Login Script
PPP Domain
PPP Account
PPP Password
Verify PPP Password
(Note: This will be supported in the future.)
•
•
•
•
•
•
Phone number of the transfer target
Login script, if applicable
Domain name of the transfer target
Login name (account) of the transfer target
Password to connect to the transfer target
Password to verify connection to the transfer
target
6.
[Save]
Saves the settings (system reboot required).
○
X
7.
[Delete]
Deletes the saved logical name (system reboot
required).
○
X
8.
[Find]
Searches for the specified computer name (if
applicable) on the network.
○
○
9.
[Test]
Searches for the specified TCP/IP address on the
network.
○
○
10.
[Home]
Returns the system to the service menu main
window.
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!77
(3)
Help screen
!
(4)
Brief description of the procedure
(a)
Enter the host name to be registered.
(b)
Enter the IP address.
(c)
Select "RAS" to connect the host via RAS (telephone circuit).
(d)
When RAS is to be used, enter the required data.
(e)
Press [SAVE] to save the entered data. The newly defined host is registered to
the Select Host pull-down menu.
4-!78
4.3.7.4.2
Directories
(1)
Purpose
Setting details of the DICOM transfer host. (This setting is required to transmit/receive
files via a LAN off line.)
(2)
Display window
!
No.
Check item
Description
LM
R
M
1.
Select Logical Name
The list of the logical names of the transfer
targets stored using this window is
displayed in the pull-down menu.
○
○
2.
Logical Name
Enter the logical name given for
convenience (effective only in this
system).
○
○
3.
FTP Host Selection
Select the host name specified on the
"Targets" page of "File Transfer".
○
○
4.
[Save]
Stores the entered data (system reboot
required).
○
X
5.
[Delete]
Deletes the stored data (system reboot
required).
○
X
6.
[Home]
Returns the system to the service menu
main window.
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!79
(3)
Help screen
!
(4)
Brief description of the procedure
(a)
Enter the logical name to be registered in the Logical Name field.
(b)
Select the host name to be associated from the FTP Host Selection pull-down
menu.
(c)
Press the [Save] button to save the settings in the system.
(d)
The newly defined logical name is registered in the Select Logical Name pull-down
menu.
4-!80
4.3.8
Service Config.
(1)
Purpose
Looking up and updating the user menu item display settings and automatic transmission
settings.
(2)
Display window
!
No.
1.
Check item
Description
Service
• Represent
ative
LM RM
○
○
○
○
Looking up and updating the name of the
nearest service representative.
• Phone No. Looking up and updating the phone number
of the nearest service representative.
2.
System Test
LM:
RM:
Note: Do not remove the check mark for
"Hide" until the System Test function
is supported.
Hide
• Check the box for "Hide" to hide System
Test in the maintenance menu.
Setting of automatic transmission for
System Test in maintenance menu
• No send
• To not perform automatic transmission
• Send event only
• To transmit the event only
• Send with full
attachment
• To transmit the results
Local mode
Remote mode
O: Effective
X: Not effective
4-!81
No.
3.
4.
Item
Description
System Maintenance
Hide • Check the box for "Hide" to hide System
Maintenance in the maintenance menu.
Setting of automatic transmission for System
Maintenance in the maintenance menu
• No send
• To not perform automatic transmission
• Send event only
• To transmit the event only
• Send with full
attachment
• To transmit the results
Image Quality Report
LM
RM
○
○
○
○
○
○
Hide • Check the box for 'Hide" to hide Image Quality
Report in the maintenance menu.
Setting of automatic transmission for Image Quality
Report in the maintenance menu
5.
• No send
• To not perform automatic transmission
• Send event only
• To transmit the event only
• Send with full
attachment
• To transmit the results
Communication
Note: Do not remove the check mark for "Hide"
until the Communication function is
supported.
Hide
• Check the box for "Hide" to hide
Communication in the maintenance menu.
Setting of automatic transmission for
Communication in the maintenance menu
• No send
• To not perform automatic transmission
• Send/Receive available
• Transmission and reception
• Receive only
• Reception only
6.
[Cancel]
Abandons all the data entered in this window and
overwrites it with the data stored in the system or in
a file.
○
○
7.
[Save]
Saves the entered data in the system.
○
○
8.
[Home]
Returns the system to the service menu main
window.
○
○
LM:
RM:
(3)
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Specify the maintenance menu show/hide setting.
(b)
Specify the automatic transmission condition for the maintenance menu.
(c)
Press [Save] to save the settings.
4-!82
4.4
4.4.1
Test Tools Function
Diagnostics
(1)
Purpose
Performing tests on diagnostic ultrasound systems.
Note: To perform automatic tests, press the Freeze switch to stop transmission and
disconnect all of the transducers from the system before entering the service
menu.
CAUTION:
Disconnect the transducers only after stopping transmission by
pressing the Freeze button. Otherwise, the transducers may be
damaged.
When the automatic tests are completed, terminate the service menu and shut down the
system.
It is recommended that the manual tests be executed first and then the automatic tests.
The manual tests cannot be started after the automatic tests.
If each board has to be tested in more detail, perform testing according to the procedure
described in section 8 "STARTING THE BOARD TESTS".
(2)
Display window
(a)
Test screen
Test Tool Diagnostic
Information
Test Tools
Diagnostic
HDD
DICOM
Probe
Front End
Maintenace
Utilities
Short Cut
Home
Date
2004/04/01 15:01
Engineer
Automatic
TI
Manual
Pass
Fail
Monitor/VCR
TX
Execute
Panel/TCS
RX
IO (RM)
RC
Speaker
BE
MIC
VP
USB
RS232C
CB
Ether
IP
Physio
!
All Select
Times
New
Ready
4-!83
Save
No.
Item
Description
LM
R
M
1.
Date
The current date is displayed
automatically. Previous test execution
dates are displayed in a pull-down menu.
○
○
2.
Engineer
Enter the name of the engineer
executing the tests.
○
○
3.
Automatic tests
• Check boxes
○
○
○
X
4.
• Select the tests to be executed.
• Result
display area
• The test result for each unit is
indicated as Pass or Fail.
• [All Select]
• Selects all the test items.
• Test count
• Any value from 1 to 100 can be
entered for the number of executions.
The default is 1.
• [Execute]
• Executes the automatic tests for each
PWB and unit the specified number of
times (system reboot required).
• [Execute and
disconnect
line]
• Executes the board test (even if the
system is disconnected from the
network) and sends the results to the
appropriate folder.
• [Cancel]
• Cancels execution of tests after
completion of the test in progress.
• Information
• Instructions, such as those for
disconnecting the transducer, are
displayed.
Manual tests
• Pass and Fail check
boxes
• Perform visual checks and enter the
results. Some test results are
checked automatically.
• [Execute]
• Displays supporting data for test
execution. Voice is output.
5.
[New]
• Acquires the current time.
○
○
6.
[Save]
• Saves the test results in the HDD.
Up to 64 files can be saved.
• When the number of files exceeds 64,
the oldest file is deleted automatically.
○
○
7.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!84
(b)
Panel/TCS test screen
When the [Execute] button is pressed, the screen below is displayed and the PANEL/
TCS test is started.
!
4-!85
No.
Item
Description
1.
ROM
The ROMs in the circuit in the panel are checked
automatically.
2.
RAM
The RAMs in the circuit in the panel are checked
automatically.
3.
Button Test 1-5
The LEDs on the switches of the panel are lit in
sequence and the peripheral circuits for the
switches are checked when the switches
corresponding to the lit LEDs are pressed. The
peripheral circuits for the following switches are
checked.
• Button Test1-----------New Patient switch
• Button Test2-----------TRANSDUCER switch
• Button Test3-----------MENU switch
• Button Test4-----------FREEZE switch
• Button Test5-----------CDI switch
4.
Track Ball Test
Move the cursor on the screen using the
trackball and check the coordinates.
The result should be entered by the engineer.
5.
STC Test
STC 1-8
Move each STC sliding VR from the minimum to
maximum and check whether the value varies
from 0 to 255.
The result should be entered by the engineer.
6.
Vol Test
Vol 1-6
Rotate the VR knobs from the minimum to
maximum and check whether the value varies
from 0 to 255.
The result should be entered by the engineer.
• Vol1 ---------------------PCG POSI VR knob
• Vol2 ---------------------ECG POSI VR knob
• Vol3 ---------------------AUDIO BALANCE VR
knob
• Vol4 ---------------------PCG GAIN VR knob
• Vol5 ---------------------ECG GAIN VR knob
• Vol6 ---------------------AUDIO volume knob
7.
TCS Test
Press the switches displayed on the touch panel
screen and check the corresponding responses.
The test result is checked automatically.
8.
Full Key Board Test
Press the keys of the keyboard and check
whether the corresponding characters are
displayed on the screen.
The result should be entered by the engineer.
4-!86
(3)
Other tests
Monitor/VCR : Check whether the images are recorded and played back correctly, and
enter the result.
(4)
Speaker
: Check whether sound is output from the speaker and enter the result.
MIC
: Check whether the voice input from the microphone is recorded and played
back correctly by the VCR, and enter the result.
USB
: Connect a USB mouse and check whether it operates correctly, and enter
the result.
Brief description of the procedure
[Automatic tests]
(a)
Confirm that all of the transducers are disconnected.
(b)
Select the board to be tested.
(c)
Enter the test counts.
(d)
Press [Execute] to start the tests.
Note: The test process may take several minutes for some boards. Pay close
attention to the message at the bottom-left of the screen.
(e)
Press [Save] to save the test results.
(f)
When the tests are completed, press [Home] to return to the service menu screen,
and then press [Exit] to close the service menu.
On the screen displayed after the service menu is closed, select Shutdown and then
turn the power OFF.
[Manual tests]
(a)
Press [Execute] to start the tests or perform the designated checks.
(b)
Enter the test results in the check boxes.
(c)
Press [Save] to save the test results.
4-!87
4.4.2
HDD
(1)
Purpose
Executing Scandisk and displaying past scandisk results.
(2)
Display window
!
No.
Item
LM RM
1.
Date
The current date is displayed automatically.
Previous test execution dates are displayed
in a pull-down menu.
Select a past execution date to display the
HDD test results for that day.
○
○
2.
Engineer
Enter the name of the engineer executing
the tests.
○
○
3.
Output
Display the results of scandisk.
○
○
4.
[Cancel]
Registers a new test execution date without
saving the test results.
○
○
5.
[Save]
Saves the test results.
○
○
6.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
(3)
Description
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Select the drive to be tested.
(b)
Press [Execute].
(c)
The test is started and the result is displayed in the text box.
(d)
Press [Save] to save the test results.
4-!88
4.4.3
DICOM
The following tests are executed according to the type of DICOM server.
(a)
Verify Network: The network settings are tested.
(b)
Storage: The file and image transmission system is tested. (This function will be
supported in the future.)
(c)
Basic print: The print settings are tested. (This function will be supported in the future.)
(1)
Purpose
Testing the connection with the remote host (executing ping).
<1>
(2)
Testing the connection with the remote DICOM AE-Title (C-ECHO).
Display window
(a)
Display window for Verify Network
!
No.
Item
Description
LM RM
1.
Host Name
Select the server to be tested.
○
○
2.
Logical Name
Select the logical name associated to the
host name.
○
○
3.
AET
Select the AE-Title to be tested.
○
○
4.
[GO]
Starts the test.
○
X
5.
[Abort]
Aborts the test.
○
X
6.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!89
(b)
Display window for Storage
!
No.
Item
Description
LM RM
1.
Extend Trace Level
<Transfer Syntax>
The trace test is performed for the DICOM
Storage application.
The Transfer Syntax pull-down menu
displays All, Implicit Little Endian (ILE),
Explicit Little Endian (ELE), Explicit Big,
Endian (EBE), JPEG Lossless (JLL), and
JPEG Lossy (JLO).
○
○
2.
Send Images
<Transfer Syntax>
Image transfer to DICOM Storage SCP is
tested.
The Transfer Syntax pull-down menu
displays All, Implicit Little Endian (ILE),
Explicit Little Endian (ELE), Explicit Big,
Endian (EBE), JPEG Lossless (JLL), and
JPEG Lossy (JLO).
The selectable items are <Image Data>,
<AET of Sender>, <Host Name>, <Logical
Name>, and <AET>.
○
○
3.
Sim. Rem. Receiver
DICOM image reception is tested.
<Host Name> and <Logical Name> can be
selected. <AET> can be specified.
○
○
4.
Sim. Loc. Receiver
Images from the specified sender are
received.
<Host Name> and <Logical Name> can be
selected. <AET> can be specified.
○
○
5.
Add. Loc. Receiver
Additional Storage SCP is started.
<AET> can be specified.
○
○
6.
[GO]
Starts the test.
○
X
7.
[Abort]
Aborts the test
○
X
8.
[Home]
Returns the system to the service menu
main window.
○
○
4-!90
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!91
(c)
Display window for Basic Print
!
No.
Item
Description
LM RM
1.
Extend Trace Level
The DICOM Basic Printer is started in trace
mode.
○
○
2.
Simulate Camera
The DICOM Reception Basic Printer
simulator is started.
○
○
3.
[GO]
Starts the test.
○
X
4.
[Abort]
Aborts the test.
○
X
5.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!92
4.5
Maintenance
4.5.1
4.5.1.1
File Maintenance
User Data Backup/Restore
(1)
Purpose
Backing up user-specific data. This function enables backing up (data saving and
restoring) of the various files created and updated by user operation. This function
allows the user-specific data to be recovered when the data file is lost due to an
incorrect operation or a problem with the magnetic disk.
The following three types of data can be backed up.
MCR
: User-defined measurement data (values set in Application Preset
Editor in Exam Type Edit in the System Setting pop-up menu)
Preset
: User-defined system presets (the preset values set in General Preset,
Peripheral System, OUTPUT, and Exam Information) and imaging
presets (imaging parameter information that can be saved for each
Exam Type by pressing the Preset switch on the touch panel)
Service Preset : Presets related to servicing (the values set using the service menu)
Note:
Removable media must be used to back up the user-specific data.
The system identification data must be stored on the media. Otherwise, the
media is ejected automatically and the data cannot be saved. Each system
requires dedicated media.
(2)
Display window
!
4-!93
No.
Item
1.
Description
Backup/Restore
• Ba
cku
p
LM
R
M
○
○
• Data is transferred from the system to
media (including a remote HDD).
• Re • Data is transferred from media
stor
(including a remote HDD) to the
e
system.
2.
Drive & Dir
Specify the media (including a remote
HDD)and directory to be used for backup
or restoration.
○
○
3.
[New Folder]
Creates a new directory
○
○
4.
[Refresh Drive List]
Initializes (clears) the area (specified in
the Drive&Dir field) to which the backup
data is to be copied.
○
○
5.
[Initialize Media (Folder)]
The medium for backup is logically
formatted using system identification data.
A medium (directory) not formatted with
this function cannot be used for backup.
○
○
6.
[Eject MO]
Ejects the medium from the MO drive.
○
○
7.
[Cancel]
Before [Go] is selected: The selections are
returned to the initial status.
During execution (after [Go] is selected):
Backup/restoration is aborted. Copying of
data is aborted after the current copy
operation is completed.
○
○
8.
[Go]
Executes Backup/Restore operation.
○
○
9.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
(3)
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Select Backup or Restore. (Only the procedure for backup is given below.)
(b)
In the Drive&Dir field, select the medium or directory to which the data is to be
backed up. To create a directory, use [Make Dir].
(c)
Initialize the directory using [Initialize (Partition)].
(d)
Select the type of data to be backed up from among MCR, Preset, and
ServicePreset.
(e)
Press [Go] to execute the backup operation.
When backup is completed successfully, a directory is created in the directory
specified above, with the date and time of backup of the data.
4-!94
4.5.1.2
File Transfer
(1)
Purpose
Transferring files concerning various data items for the system using FTP.
The following "sources" are provided.
(2)
(a)
Image
: To transfer images from the system
(b)
File
: To transfer files from the system
(c)
Download List : To view the list of files transferred from the remote device to the
system
(d)
Extract List
: To view the list of files in the transfer directory in the system
Display window
(a)
Display window for Image
!
4-!95
No.
1.
Item
Description
LM
RM
○
○
Action
• Transfer Online
• Specify the patient name, study, and series to
display thumbnail images. The displayed
images are transferred to a remote terminal.
• Extract Only
• Specify the patient name, study, and series to
display thumbnail images. The specified file is
copied to the transfer directory.
• Display Header
• Specify the patient name, study, and series to
display thumbnail images.
2.
Patient
The list of patient numbers is displayed in the pulldown menu.
○
○
3.
Study
The list of study numbers is displayed in the pulldown menu.
○
○
4.
Series
The list of series numbers is displayed in the pulldown menu.
○
○
5.
Selected Images
The list of selected images is displayed.
○
○
6.
Target
Select the transfer destination remote terminal.
○
○
7.
Show Thumb Nails
Select whether or not thumbnail display is to be
enabled.
○
○
8.
[Delete]
Deletes the selected images.
○
○
9.
[Reset]
Initializes image selection.
○
○
10.
[Go]
Transfers the selected images.
○
○
LM:
RM:
(b)
Local mode
Remote mode
O: Effective
X: Not effective
Display window for File
The menu screen differs depending on the selected action.
No.
LM:
RM:
Item
Description
L
M
R
M
1.
Transfer Online
Specify the patient name, study, and
series for thumbnail image display. The
displayed images can be transferred
directly to a remote terminal (operating
terminal).
-
-
2.
Extract Only
Specify the patient name, study, and
series for thumbnail image display. The
files are copied to the transfer directory.
-
-
3.
Download
The specified files are downloaded
from a remote device to the system.
-
-
Local mode
Remote mode
4-!96
(b)-1 Menu screen for Transfer Online
!
No.
1.
Item
Description
File Type
• Any file
LM
R
M
○
○
• Any of the files on the system disk are
transferred to a remote terminal.
• Most common logfiles
(Note:
This will
be supported in
the future.)
• Service related logfiles
(Note:
This will
be supported in
the future.)
• All available logfiles
(Note:
This will
be supported in
the future.)
2.
Filter
File types are filtered.
○
○
3.
[Refresh]
Refreshes the display of the directory
information.
○
○
4.
[Go]
Starts the file transfer operation.
○
○
5.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!97
(b)-2 Menu screen for Download
!
No.
1.
Item
File Type
•
LM:
RM:
Description
L
M
R
M
○
○
Any file • Any of the files on a remote terminal
are downloaded to the system.
2.
Source
Select the remote terminal.
○
○
3.
[Refresh]
Refreshes the display of the download
directory information.
○
○
4.
[Go]
Starts the download operation.
○
○
5.
[Home]
Returns the system to the service menu
main window.
○
○
Local mode
Remote mode
O: Effective
X: Not effective
4-!98
(b)-3 Menu screen for Extract Only
!
No.
1.
Item
Description
File Type
• Any file
L
M
R
M
○
○
• Any of the files on the system disk
are copied to the transfer directory.
• Most common
logfiles
(Note:
This
will be
supported in
the future.)
• Service related
logfiles
(Note:
This
will be
supported in
the future.)
• All available logfiles
(Note:
This
will be
supported in
the future.)
LM:
RM:
2.
[Refresh]
Refreshes the display of the directory
information.
○
○
3.
[Go]
Copies the file to the transfer directory.
○
○
4.
[Home]
Returns the system to the service menu
main window.
○
○
Local mode
Remote mode
O: Effective
X: Not effective
4-!99
(c)
Display window for Download List
!
No.
LM:
RM:
Item
Description
L
M
R
M
○
○
1.
Action
2.
Site Directory
The location of the directory in the
system is displayed.
○
○
3.
Filter
File types are filtered.
○
○
4.
[Refresh]
Refreshes the display of the directory
where the downloaded files are to be
stored.
○
○
5.
[Delete]
Deletes the downloaded files.
○
○
6.
[Home]
Returns the system to the service menu
main window.
○
○
Local mode
Remote mode
• Show/ The list of files is displayed. Or the files
Delete are deleted.
O: Effective
X: Not effective
4-!100
(d)
Display window for Extract List
!
No.
1.
LM:
RM:
Item
Description
Action
• Transfer
Online
• The files in the transfer directory are
transferred.
• Show/
Delete
• The list of files in the transfer
directory is displayed. Or the files in
the transfer directory are deleted.
L
M
R
M
○
○
2.
Filter
File types are filtered.
○
○
3.
[Refresh]
Refreshes the display of the transfer
directory information.
○
○
4.
[Go]
Executes the file transfer operation.
○
○
5.
[Home]
Returns the system to the service menu
main window.
○
○
Local mode
Remote mode
O: Effective
X: Not effective
4-!101
4.5.2
4.5.2.1
Log
Original Log Viewer
(1)
Purpose
Enables the results of maintenance operations, such as System Test, Diagnostic,
System Maintenance, and Image Quality Report in the maintenance menu, to be
viewed.
(2)
Display window
(a)
Initial window
!
4-!102
(b)
System Test Log window
!
(c)
Diagnostic Log window
!
4-!103
(d)
System Maintenance Log window
!
(e)
Image Quality Report Log window
!
4-!104
(f)
Operation Log window
!
4-!105
4.5.2.2
Event Viewer
(1)
Purpose
Viewing the event log through simplified operation.
(2)
Display window
!
4-!106
No.
1.
Item
Description
Domain
• Application
• Security
• System
2.
Severity
•
•
•
•
Error
Warning
Information
All
LM
RM
Select the type of log.
• Application log
• Security log
• System log
○
○
Select the severity (error level).
• Errors are displayed.
• Warnings are displayed.
• Information is displayed.
• All levels of messages are displayed.
○
○
3.
Searchorder
• Newest first
• Oldest first
Select the display order.
• The newest data is displayed first.
• The oldest data is displayed first.
○
○
4.
Message Limit
Specify the maximum number of messages to be
displayed.
○
○
5.
Time Range
Specify the date range for display of the messages.
• Start date
• End date
○
○
• From
• To
6.
Time
Check this box to narrow down the range to
particular hours on the specified date.
○
○
7.
[Exit]
Closes the window.
○
○
8.
[Abort]
Aborts the search operation.
○
○
9.
[Go]
Starts the search operation.
○
○
LM:
RM:
(3)
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Select the log type.
(b)
Select the desired message in the search result list.
(c)
Details are displayed for the selected message.
4-!107
4.5.3
Report
(1)
Purpose
Enables electronic data from users and service engineers at the user site to be viewed.
(2)
Display window
!
No.
Item
Description
LM RM
1.
Date/Time
Date and time of data creation
○
○
2.
Subject/Directory
Data title
○
○
3.
From
Name of writer
○
○
4.
Category
Data category
○
○
5.
Subject
Title given by responding service engineer
○
○
6.
Name
Name of responding service engineer
○
○
7.
[Cancel]
All the data entered in this window is
abandoned.
○
○
8.
[Save]
The entered data is transferred to the
system at the user site and saved there.
The corresponding data that was used
previously is deleted.
○
○
9.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!108
4.5.4
Service History
(1)
Purpose
Creating the service history.
(2)
Display window
!
No.
Item
Description
LM RM
1.
Engineer
Enter name of the engineer in charge of
service.
○
○
2.
Date
If "New" is specified, the current date is
automatically read and displayed.
If a past date is specified, the data
corresponding to the specified date is
displayed.
○
○
3.
Type of Maintenance
Select either proactive maintenance or
reactive maintenance.
○
○
4.
Contents
Check the boxes for the target service
items.
○
○
5.
Frequency of
Phenomenon
Select the frequency of recurrence of the
problem.
○
○
6.
Current Condition
Select the servicing status.
○
○
7.
Working time
Enter the actual time required for the work.
○
○
8.
Memo
Enter the work contents and comments.
○
○
9.
[Cancel]
If "New" is specified, all the data entered in
the window is deleted.
○
○
10.
[New]
Used to enter the data for the work
performed on the current date and to
initialize other data.
○
○
11.
[Save]
Transfers and saves the entered data to the
user system.
○
○
12.
[Home]
Returns the system to the service menu
main window.
○
○
4-!109
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!110
(3)
4.6
4.6.1
Brief description of the procedures
(a)
Enter the name of the service engineer.
(b)
Check the boxes in the Maintenance Work section according to the current servicing
work.
(c)
Press [Save] to save the data in the Aplio system
Utilities
Command Line
(1)
Purpose
Executing a command line (DOS command) and displaying results of command execution.
(2)
Display window
!
No.
Item
Description
LM RM
1.
Command Line
Enter the command to be executed.
○
○
2.
Output
The command execution results are
displayed.
○
○
3.
[Explorer]
Displays Explorer.
○
X
4.
[Control Panel]
Displays the Control Panel.
○
X
5.
[MO Eject]
Ejects the MO from the drive.
○
○
6.
[Go]
Executes the entered command.
○
○
7.
[Abort]
Aborts the command being executed.
○
○
8.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!111
4.6.2
Stationing application
(1)
Purpose
Enables the α protocol and the card reader to be turned ON/OFF.
(2)
Display window
!
No.
1.
Item
Description
LM RM
Alpha Protocol
Specify whether α protocol is to be used or
not.
○
○
Card Reader
Specify whether the card reader is to be
used or not.
○
○
3.
[Cancel]
Initializes the settings.
○
○
4.
[SAVE]
Saves the settings made in this window.
○
○
5.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-101!A
*
4.6.3
Remote Application
(1)
Purpose
Opening a window on the diagnostic ultrasound system to display messages.
(2)
Display window
!
No.
Item
LM RM
1.
Output
Enter the message to be displayed on the
diagnostic ultrasound system.
○
○
2.
[Cancel]
Clears the entered message.
○
○
3.
[Display]
Displays the message on the diagnostic
ultrasound system.
○
○
4.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
(3)
Description
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedures
(a)
Enter the message to be displayed on the diagnostic ultrasound system.
(b)
Press [Display].
(c)
The entered message is displayed on the diagnostic ultrasound system.
4-!102
4.7
4.7.1
Shortcuts
Smart Backup
(1)
Purpose
Backing up the service information through simplified operation.
(2)
Display window
!
4-!103
No.
Item
LM RM
1.
Backup/Restore
Backup : System to media (including
remote HDD)
Restore: Media (including remote HDD) to
system
○
○
2.
Drive & Dir
Specify the media (including remote HDD)
and directory to be used for backup or
restore.
○
○
3.
[New Folder]
Creates a new directory.
○
○
4.
[Refresh Drive List]
Initializes (clears) the area (specified in the
Drive&Dir field) to which the backup data is
to be copied.
○
○
5.
[Initialize Media (Folder)]
The medium for backup is logically
formatted using system identification data.
A medium (directory) that is not formatted
with this function cannot be used for
backup.
○
○
6.
[Eject MO]
Ejects the medium from the MO drive.
○
○
7.
[Cancel]
Before [Go] is selected: The selections are
returned to the initial status.
During execution (after [Go] is selected):
Backup /Restore is aborted. Copying of
data is aborted after the current copy
operation is completed.
○
○
8.
[Go]
Executes the Backup /Restore operation.
○
○
LM:
RM:
(3)
Description
Local mode
Remote mode
O: Effective
X: Not effective
Brief description of the procedure
(a)
Select Backup.
(b)
In the Drive&Dir field, select the medium or directory to which the data is to be
backed up. To create a directory, use [Make Dir].
(c)
Initialize the directory using [Initialize (Partition)].
(d)
Select the type of data to be backed up from among MCR, Preset, and
ServicePreset.
(e)
Press [Go] to execute the backup operation.
When backup is completed successfully, a directory is created in the directory
specified above, with the date and time of backup of the data.
4-!104
4.7.2
Software Upgrade
(1)
Purpose
The software upgrade procedure is displayed and the automatic login process is
terminated.
(2)
Display window
!
No.
Item
Description
LM RM
1.
Software Upgrade
The procedure for upgrading the software is
displayed.
○
○
2.
[Go]
The automatic login process is terminated
(system reboot required).
○
○
3.
[Home]
Returns the system to the service menu
main window.
○
○
LM:
RM:
Local mode
Remote mode
O: Effective
X: Not effective
4-!105
E