Uploaded by Sahir Khan

Rotors-Manual

advertisement
Rotor Fitment and Details
Centurion
Scientific Limited
Welcome
It is of p
a
import ramount
read thisance you
doc
before u ument
se
to our Rotor Fitment, balancing,
care and safety manual
Centrifuges can be dangerous if rotors are not correctly fitted, samples
are not correctly loaded or poor samples tubes are used.
Rotor fitment is from Page 8-41. Please look up the rotor to be used to
go to the relevant pages. Read carefully and follow the instructions.
We recommend:
● Quality tubes are used
● Y
ou balance within
1 gram
● Y
ou balance the rotor
correctly.
● Y
ou always fit the lid
supplied
● W
e recommend that
you check the rotor for
cracks reguarly.
!
2
Tubes can break or rupture causing
damage or sample failure. We know
how much time & work it may take
to get to this point to be ruined by a
cheap tube. Although our imbalance
system is sophisticated and will shut
down if exceeded. Centrifugation is
about seperation and good balancing
gives better separations.
Always place balanced tubes opposite
to each other, as shown, in the pictures.
Lids are used to prevent aerosol effects
and reduce air flow and therefore noise.
If a crack is found. DO NOT USE.
Rotors are under a constant strain and need to be kept clean and checked for
strain.
DO NOT exceed the life of the rotor 7 years maximum.
For samples of an unknown nature we recommend you check before centrifugation
by using the Chemical Resistance Table, page 42 onward.
We do not recommmend the use of strong acids, strong Alkalis or strong
radioactive samples, these can damage the rotors and the Centrifuge.
www.centurionscientificglobal.com
Index
Page 4:
Balance. Haematocrit and Micro
Page 5:
Balance fixed angle
Page 6:
Balance swing outs
Page 7:
Balance Cytology & Cell wash
Page 8 - 9:
BRK5401
Page 10 - 11:
BRK5424. BRK5436. BRK5448. BRK5494
Page 12 - 13:
BRK5407. 5408
Page 14 - 15:
BRK5412. BRK5406
Page 16 - 17:
BRK5308. BRK5324. BRK5100
Page 18 - 19:
BRK5431. BRK5434
Page 20 - 21:
BRK5508.
Page 22 - 23:
BRK5505.
Page 24 - 25:
BRK5510. New BRK1000, with 100ml bucket
Page 26 - 27:
BRK5510. New BRK1000, with 250ml bucket
Page 28 - 29:
BRK5553. New BRK3000 & BRK4000 (Same fitment)
Page 30 - 31:
BRK5540
Page 32 - 33:
BRK5206. BRK5212
Page 34 - 35:
BRK5210. BRK5208. BRK5224
Page 36 - 37:
BRK5256
Page 38 - 39:
4420. 4430. 4460
Page 40 - 41:
5450. C washer rotor
Page 42
Cleaning
Page 43:
Nomograph
Page 44 - 54:
Page 55:
Page 56 - 59:
Chemical Resistance table
Contact us
Certificates
Rotor Fitment and Details
3
Rotor Balance
Haematocrit Rotor
!
✔
✘
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit the lid supplied
CHECK the Rim seal gasket and replace regularly (Part number 8011)
CHECK the rotor screw fixing tightness regularly
Micro Rotors
!
4
✔
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit the lid supplied
CHECK the two top seals and replace regularly (Part number 8044)
CHECK the rotor screw fixing tightness regularly
www.centurionscientificglobal.com
✘
Small fixed angle Rotors
!
✔
✘
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
CHECK The rotor screw fixing tightness regularly
Large fixed angle Rotors
!
✔
✘
✘
✔
✘
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
CHECK The rotor screw fixing tightness regularly
Rotor Fitment and Details
5
Small swing out Rotor
!
✔
✘
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit ALL the Tube Holders and any supports
CHECK The rotor screw fixing tightness regularly
Medium – Large Swing out Rotors
!
✔
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit ALL 4 Buckets and 4 Adaptors supplied
CHECK the rotor screw fixing tightness regularly
Microtitor Rotor
!
6
✔
✘
ALWAYS load your plates opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit ALL 2 Buckets supplied
CHECK the rotor screw fixing tightness regularly
www.centurionscientificglobal.com
✘
Cytology Rotors
!
✔
✘
✘
ALWAYS load your holders opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit the lid supplied
Check the two top Gaskets and replace regularly (Part Number 8055)
CHECK the rotor screw fixing tightness regularly
Manual Cell Washing Rotor
!
✔
✘
✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
Regularly smear a small amount of Silicon grease over the O rings
Check the inner and outer O rings and replace regularly (Part Number 8066)
CHECK the rotor screw fixing tightness regularly
Rotor Fitment and Details
7
Rotor fitment
5
1
2 34
6
7
1
2
Place the rotor on
the shaft of the
motor.
Locate threaded
fastener (Item
2) into the three
interlocking pin
holes on the rotor.
BRK5401
1.
2.
3.
4.
5.
6.
7.
Rotor
Threaded Fastener
M6 Spring Washer
Rotor Screw (M6 x 20mm)
Rotor Lid
Rotor Lid Knob
Allen Key
6 Place samples in
as required.
3 Place M6 Spring
7 Place rotor lid
5 Hand tighten with
8 Hand tighten lid
washer (Item3)
onto rotor screw
(Item 4).
screw
4 Insert
and washer into
central hole of
threaded fastener.
Allen Key (Item 7)
(Item 5) on
threaded fastener.
using rotor knob
(Item 6).
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED ON A
REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid.
8
www.centurionscientificglobal.com
Rotor fitment
Rotor
BRK5401
Rotor type
24 x capillary/12 x 2ml
Tube size max
2 x 75mm/11 x 40mm
Minimum speed Rpm
500
Maximum speed Rpm
12,000
Maximum Rcf (G)
13,500
Radius max cms
8.4
Sample tube angle °
0/60
Acceleration time (secs)
25
Deceleration time (secs)
25
Autoclavable (frequecncy)
121°C (10)
Accessories
Part Number
Description
8012
Plain Capillary tubes pack 1000
8013
Heparanised Capillary tubes pack 1000
8003 PCV reader
8014 Sealing Clay
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
9
Rotor fitment
BRK5424. BRK5436.
BRK5448. BRK5494
1
4
2
3
5
1 Locate threaded
fastener (Item
3) into the three
interlocking pin
holes on the rotor.
2 Place M6 Spring
washer (Item 2)
onto rotor screw
(Item 1)
3
Insert screw
and washer into
central hole of
threaded fastener.
1. Rotor Screw (M6 x 20mm)
2. M6 Spring Washer
3. Threaded Fastener
4. Torque Key
5. Rotor Lid
Hand tighten with
4 Torque Key
(Item 4)
5
Always fit the lid
provided
6
To fasten the
rotor lid, Place
rotor lid (Item
5) on threaded
fastener hand
tighten using the
rotor knob.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
10
www.centurionscientificglobal.com
Rotor Details
Microtube rotors
Rotor
BRK5424
BRK5436
BRK5448
BRK5494
Rotor type
24 x 2.2ml
36 x 0.5ml
48 x 0.2 to 0.4ml
4 x Pcr Strips
Tube size max
11 x 50mm
8 x 30mm
6 x 40mm
8 per strip
Minimum speed Rpm
500
500
500
500
Maximum speed Rpm
15,000
15,000
15,000
15,000
Maximum rcf (G)
22,000
22,000
22,000
22,000
Radius max cms
8.75cms
8.75cms
8.75cms
8.75cms
Sample tube angle °
45
45
45
45
Acceleration time (secs)
20
20
20
20
Deceleration time (secs)
20
20
20
20
Autoclavable (frequecncy)
121°C (10)
121°C (10)
121°C (10)
121°C (10)
Reducers
Rotor
BRK5424
Part number
RS04 (0.2- 0.4ml)
Tube size max
6 x 30mm
Part number
RS05 (0.5ml)
Tube size max
8 x 30mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
11
Rotor fitment
BRK5407. BRK5408
2
3
1
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
Place the rotor
3 fixing screw
(Item2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(Item 3).
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
12
www.centurionscientificglobal.com
Rotor Details
Rotor
BRK5407
BRK5408
Rotor type
4 x 15ml & 4 x 10ml
8 x 15ml tube
Tube size max
17 x 120mm & 13 x 110mm
17 x 120mm
Minimum speed Rpm
500
500
Maximum speed Rpm
8,000
6,000
Maximum Rcf (G)
6,100
3,500
Radius max cms
8.6
8.5
Sample tube angle °
30
30
Acceleration time (secs)
30
30
Deceleration time (secs)
30
30
Autoclavable (frequecncy)
121°C (10)
121°C (10)
Reducers
Rotor
BRK5407 / BRK5408
Part number
RM05 (5ml)
Tube size max
13 x 80mm
Part number
RM10 (10ml)
Tube size max
16 x 100mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
13
Rotor fitment
BRK5412. BRK5406
1
1. M6 Rotor Screw
2. M6 Washer
2
1 Place the rotor on
the shaft of the
rotor.
2 Place the M6
Washer (item 2)
over the centre
hole on the rotor.
the M6
3 Place
rotor fixing (item
1) into the central
hole of the rotor.
4
Hand tighten the
M6 rotor fixing.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
14
www.centurionscientificglobal.com
Rotor Details
Rotor
BRK5412
BRK5406
Rotor type
12 x 15ml
6 x 50ml
Tube size max
17 x 120mm
30 x 120mm
Minimum speed Rpm
500
500
Maximum speed Rpm
6,000
6,000
Maximum rcf (G)
3,800
3,800
Radius max cms
9.5
9.5
Sample tube angle °
30
30
Acceleration time (secs)
25
25
Deceleration time (secs)
25
25
Autoclavable (frequecncy)
121°C (10)
121°C (10)
Reducers
Rotor
BRK5412
BRK5406
Part number
RM05 (5ml)
RM15 (15ml)
Tube size max
13 x 80mm
17 x 120mm
Part number
RM10 (10ml)
RM25 (25ml)
Tube size max
16 x 100mm
25 x 100mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
15
Rotor fitment
BRK5308. BRK5324.
BRK5100
1
2
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
1. M6 Washer
2. M6 Handwheel
Place the rotor
3 fixing screw
(Item2) into the
central hole of the
rotor.
4
Hand tighten with
Allen Key (Item 3).
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
16
www.centurionscientificglobal.com
Rotor Details
Rotors 6000 RPM
BRK5324
Rotor
BRK5308
BRK5100
Rotor type
24 x 15ml
8 x 50ml
6 x 100ml
Size max
17 x 120mm
30 x 120mm
45 x 125mm
Minimum speed Rpm
500
500
500
Maximum speed Rpm
6,000
6,000
6,000
Maximum Rcf (G)
4,800
4,800
4,800
Radius max cms
12
12
12
Sample tube angle °
30
30
30
Acceleration time (secs)
35
35
35
Deceleration time (secs)
35
35
35
Autoclavable (frequency)
121°C (10)
121°C (10)
121°C (10)
4°C
4°C
Refrigerated Centrifuges Only
Minimum Temperature
4°C
At maximum speed (relative to room temperature at 23°C)
Pack of 4
Reducers
Pack of 4
Rotor
BRK5324
BRK5308
BRK5100
Part number
RM05 (5ml)
RM15 (15ml)
RL10 (10ml)
Tube size max
13 x 80mm
17 x 120mm
16 x 100mm
Part number
RM10 (10ml)
RM25 (25ml)
RL25 (25ml)
Tube size max
16 x 100mm
25 x 100mm
25 x 100mm
RL50 (50ml)
35 x 110mm
RL85 (85ml)
39 x 110mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
17
Rotor fitment
BRK5431. BRK5434
1
2
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
1. M6 Handwheel
2. M6 Washer
Place the M6
3 rotor fixing knob
(Item2) into the
central hole of the
rotor.
4
Hand tighten M6
rotor fixing knob.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
18
www.centurionscientificglobal.com
Rotor Details
Rotors 4000 RPM
Rotor
BRK5431
BRK5434
Rotor type
24 x 15ml
8 x 50ml
Tube size max
17 x 120mm
30 x 120mm
Minimum speed Rpm
500
500
Maximum speed Rpm
4,000
4,000
Maximum rcf (G)
2,200
2,200
Radius max cms
12.6
12.6
Sample tube angle °
30
30
Acceleration time (secs)
35
35
Deceleration time (secs)
35
35
Autoclavable (frequecncy)
121°C (10)
121°C (10)
Reducers
Rotor
BRK5431
BRK5434
Part number
RM05 (5ml)
RM15 (15ml)
Tube size max
13 x 80mm
17 x 120mm
Part number
RM10 (10ml)
RM25 (25ml)
Tube size max
16 x 100mm
25 x 100mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
19
Rotor fitment
BRK5508, S, M, L
3
2
1
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
Place the M6
3 rotor fixing screw
(Item 2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(item 3). Ensure
all buckets are
fully inserted
before use.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
20
www.centurionscientificglobal.com
Swing out Rotor - 8 x 5ml, 8 x 10ml & 8 x 15ml
Rotor
BRK5508S
BRK5508M
BRK5508L
Rotor type
8 x 5ml Swing Out
8 x 10ml
8 x 15ml
Tube size max
13 x 75mm max
15 x 115mm
17 x 125mm
Minimum Speed Rpm
500 Rpm
500 Rpm
500 Rpm
Maximum Speed Rpm
4,000 Rpm
4.000 Rpm
4.000 Rpm
Maximum Rcf (G)
1500
2200
2,600
Radius max cms
8.5
12.4
14.6
Sample tube angle (°)
0
0
0
Acceleration time (secs)
20
20
20
Deceleration time (secs)
20
20
20
Autoclavable (frequency)
121°C (20)
121°C (20)
121°C (20)
Rotor Fitment and Details
21
Rotor fitment
BRK5505
3
2
1. M6 Washer
2. Rotor Screw
M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
1
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
3 Place the M6
rotor fixing screw
(Item 2) into the
central hole of the
rotor.
Hand tighten with
4 Torque Key
(item 3).
5
Ensure all four
buckets are fully
inserted before
use.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
22
www.centurionscientificglobal.com
Adaptors for BRK5505
Capacity
Rotor Details
Rotors
BRK5505
Rotor type
4 x 100ml
Tube size max
45 x 100mm
Minimum speed Rpm
500
Maximum speed Rpm
4,000
Maximum rcf (G)
2,200
Radius max cms
12.5
Sample tube angle °
0
Acceleration time (secs)
25
Deceleration time (secs)
25
Autoclavable (frequecncy)
121°C (20)
Sealed lids available B5419
PLEASE ENSURE THE ROTOR IS
CHECKED REGULARLY FOR CRACKS
OR DEFECTS
Size
Part No.
Tubes per
rotor
Tube type: Micro with cap Shape: point
0.5m
8x20
AM505
36
1.5ml
11x38
AM520
24
2.0ml
11x38
AM520
24
0.2ml
6x20
AM502
36
0.4ml
6x30
AM504
36
Tube type: Plain no cap Shajje: round
1ml
6x45
AR501
36
3ml
10x60
AR503
28
5ml
12x75
AR505*
28
6ml
12x82
AR505*
28
7ml
12x100
AR507*
28
9/10ml
14x100
AR509*
24
15ml
17x100
AR515*
8
25ml
24x100 AR525*
8
50ml
34x100 AR550*
4
100ml
45x100 AR5100*
4
150ml
52x100 N/A
250ml
62x100 N/A
Tube type: Falcon with cap Shape: point
15ml
17x120
AF515*
8
50ml
29x115
AF550*
4
175ml
61x118
N/A
15ml
17x120
N/A
Tube type: Corning with cap* Shape: point
250ml
60x172
See K242/R
500ml
98x148 See K242/R
Tube type: Falcon with cap Shape: square
12ml
17x100
AFS512
8
25ml
25x90
AFS525
4
30ml
25x110
AFS530
4
50ml
29x115
AFS550*
4
15ml
17x120
N/A
N/A
Tube type: Nalgene/Oakridge Shape: round
10ml
16x80
ANO510*
8
30ml
26x95
ANO530*
4
50ml
29x107
ANO550
4
100ml
38x106
ANO585
4
Tube type: Nalgene/Oakridge* Shape: flat
250ml
62x130
See K241/R
750ml
98x153
See K243/R
Tube type: Monovette Shape: square
1.1-1.4ml
8x82
AM5014
24
2.7-3ml
11x82
AM503
24
2.6-2.9ml 13x81
AM529*
24
4.5-5ml
11x108
AM503*
24
7.5-8.2ml 13x106
AM579*
20
4.5-5ml
15x92
AM550*
20
9-10ml
16x108
AM590*
20
Tube type: Vacutainer Shape: round
1.6-5ml
13x75
AV516*
24
4-7ml
13x100
AV550*
16
8.5-10ml
16x100
AV580*
12
Note: - We can not be held responsible if the tube manufacturers change the design and/or size of the tube.*
Limited capacity with sealed bucket lids (B5419 and B5319)
Rotor Fitment and Details
23
Rotor fitment
BRK5510. New BRK1000
3
2
1. M6 Washer
2. Rotor Screw
M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
1
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
3 Place the M6
rotor fixing screw
(Item 2) into the
central hole of the
rotor.
Hand tighten with
4 Torque Key
(item 3).
5
Ensure all four
buckets are fully
inserted before
use.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
24
www.centurionscientificglobal.com
Swing Out Rotor
- 400ml max
Adaptors for Swing out rotors
Capacity
Size
To fit buckets B5010
Tubes per
rotor
Part No.
Tube type: Micro with cap Shape: point
0.5m
8x20
AM405
36
1.5ml
11x38
AM420
24
2.0ml
11x38
AM420
24
0.2ml
6x20
AM402
36
0.4ml
6x30
AM404
36
BRK1000 Swing out Rotor
Shown with B5010 Buckets and
sealed lids B5419
Rotor / buckets
BRK1010
Tube size max
45 x 100mm
Minimum speed
500Rpm
Maximum speed
4000Rpm
Maximum Rcf (G)
2200
Radius max
12.5cm
Tube angle
0 degree
Acceleration time
25 seconds
Deceleration time
25 seconds
Autoclavable (times)
121°C (20)
Minimum Temperature 4°C (at 23°C
ambient)
B5010 bucket (set 4) REQUIRED
100ml max per bucket
Tube type: Plain no cap Shape: round
1ml
6x45
AR401
36
3ml
10x60
AR403
28
5ml
12x75
AR405*
28
6ml
12x82
AR405*
28
7ml
12x100
AR407*
28
9/10ml
14x100
AR409*
24
15ml
17x100
AR415*
8
25ml
24x100 AR425*
8
50ml
34x100 AR450*
4
100ml
45x100 AR4100*
4
150ml
52x100 N/A
250ml
62x100 N/A
Tube type: Falcon with cap Shape: point
15ml
17x120
AF415*
8
50ml
29x115
AF450*
4
175ml
61x118
N/A
15ml
17x120
N/A
Tube type: Corning with cap* Shape: point
250ml
60x172
See K242/R
500ml
98x148 See K242/R
Tube type: Falcon with cap Shape: square
12ml
17x100
AFS412
8
25ml
25x90
AFS425
4
30ml
25x110
AFS430
4
50ml
29x115
AFS450*
4
15ml
17x120
N/A
N/A
Tube type: Nalgene/Oakridge Shape: round
10ml
16x80
ANO410*
8
30ml
26x95
ANO430*
4
50ml
29x107
ANO450
4
100ml
38x106
ANO485
4
Tube type: Nalgene/Oakridge* Shape: flat
250ml
62x130
See K241/R
750ml
98x153
See K243/R
Tube type: Monovette Shape: square
1.1-1.4ml
8x82
AM4014
24
2.7-3ml
11x82
AM403
24
2.6-2.9ml 13x81
AM429
24
4.5-5ml
11x108
AM403*
24
7.5-8.2ml 13x106
AM479*
24
4.5-5ml
15x92
AM450*
20
9-10ml
16x108
AM490*
20
Tube type: Vacutainer Shape: round
1.6-5ml
13x75
AV416*
24
4-7ml
13x100
AV450*
16
8.5-10ml
16x100
AV480*
12
Note: - We can not be held responsible if the tube manufacturers change the design and/or
size of the tube.* Limited capacity with sealed bucket lids (B5419 and B5319)
25
Rotor fitment
BRK5510. New BRK1000
3
2
1. M6 Washer
2. Rotor Screw
M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
1
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
3 Place the M6
rotor fixing screw
(Item 2) into the
central hole of the
rotor.
Hand tighten with
4 Torque Key
(item 3).
5
Ensure all four
buckets are fully
inserted before
use.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
26
www.centurionscientificglobal.com
Adaptors for Swing out rotors
Swing Out Rotor
1 Litre Max
Capacity
Size
To fit buckets B5250
Tubes per
rotor
Part No.
Tube type: Micro with cap Shape: point
0.5m
8x20
AM605
40
1.5ml
11x38
AM620
28
2.0ml
11x38
AM620
28
0.2ml
6x20
AM602
48
0.4ml
6x30
AM604
48
Tube type: Plain no cap Shape: round
1ml
6x45
AR601
36
3ml
10x60
AR603
28
5ml
12x75
AR605
28
6ml
12x82
AR605
28
7ml
12x100
AR607
28
9/10ml
14x100
AR609*
24
15ml
17x100
AR615*
28
25ml
24x100 AR625
8
50ml
34x100 AR650
4
100ml
45x100 AR6100
4
150ml
52x100 AR601
250ml
62x100 BUCKET 4
BRK1000 Swing out Rotor
Shown with B5250 Buckets and
sealed lids B5419
Rotor / buckets
BRK1025
Tube size max
62 x 100mm
Minimum speed
500Rpm
Maximum speed
4000Rpm
Maximum Rcf(G)
2600
Radius max
14.6cm
Tube angle
0 degree
Acceleration time
25 seconds
Deceleration time
25 seconds
Autoclavable (times)
121°C (20)
Minimum Temperature 4°C (at 23C
ambient)
B5250 bucket (set 4) REQUIRED
250ml max per Bucket
Set of 4
Tube type: Falcon with cap Shape: point
15ml
17x120
AF615*
16
50ml
29x115
AF650*
4
175ml
61x118
AF6175*
4
15ml
17x120
N/A
4
Tube type: Corning with cap* Shape: point
250ml
60x172
See K242/R
500ml
98x148 See K242/R
Tube type: Falcon with cap Shape: square
12ml
17x100
AFS612*
16
25ml
25x90
AFS625*
8
30ml
25x110
AFS630
8
50ml
29x115
AFS650
4
15ml
17x120
AFS614
4
Tube type: Nalgene/Oakridge Shape: round
10ml
16x80
ANO610
16
30ml
26x95
ANO630
4
50ml
29x107
ANO650
4
100ml
38x106
ANO685
4
Tube type: Nalgene/Oakridge* Shape: flat
250ml
62x130
Buckets 5250 4
750ml
98x153
See K243/R
Tube type: Monovette Shape: square
1.1-1.4ml
8x82
AM6014
40
2.7-3ml
11x82
AM603
40
2.6-2.9ml 13x81
AM629
40
4.5-5ml
11x108
AM603*
40
7.5-8.2ml 13x106
AM679*
40
4.5-5ml
15x92
AM650
28
9-10ml
16x108
AM690*
28
Tube type: Vacutainer Shape: round
1.6-5ml
13x75
AV616
40
4-7ml
13x100
AV650
40
8.5-10ml
16x100
AV680
28
Note: - We can not be held responsible if the tube manufacturers change the design and/or
size of the tube.* Limited capacity with sealed bucket lids (B5419 and B5319)
27
Rotor fitment
3
BRK5553. New BRK3000 &
BRK4000 (Same fitment)
2
1
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
Place the M6
3 rotor fixing screw
(Item 2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(item 3). Ensure
all four buckets
are fully inserted
before use.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
28
www.centurionscientificglobal.com
Adaptors for 5520/5553
Capacity
Rotor Details
Illustrative Purposes
Rotors
BRK5553
BucketsB3000 (set 4 required)
Rotor type
4 x 750ml
Tube size max
98 x 170mm
Minimum Speed Rpm
500
Maximum Speed Rpm
4,000
Maximum Rcf (G)
3,600
Radius max cms
20.5
Sample tube angle (°)
0
Acceleration time (secs)
45
Deceleration time (secs)
45
Autoclaveable (frequency) 121°C (20)
Rotors
BRK5520
Buckets
Rotor type
Tube size max
Minimum Speed Rpm
Maximum Speed Rpm
Maximum Rcf (G)
Radius max cms
Sample tube angle (°)
Acceleration time (secs)
Deceleration time (secs)
Autoclaveable (frequency)
B2000 (set 4 required)
4 x 500ml
98 x 155mm
500
4,000
3,400
19.5
0
45
45
121°C (20)
Refrigerated Centrifuges Only
Minimum Temperature
4°C
At maximum speed (relative to room temperature
at 23°C)
Sealed lids available B5319
Size
Part No.
Tubes per
rotor
Tube type: Micro with cap Shape: point
0.5m
8x20
AM805
120
1.5ml
11x38
AM820
84
2.0ml
11x38
AM820
84
0.2ml
6x20
AM802
168
0.4ml
6x30
AM804
168
Tube type: Plain no cap Shape: round
1ml
6x45
AR801
168
3ml
10x60
AR803
96
5ml
12x75
AR805
96
6ml
12x82
AR805
96
7ml
12x100
AR807
96
9/10ml
14x100
AR809
84
15ml
17x100
AR815
48
25ml
24x100 AR825
28
50ml
34x100 AR850
16
100ml
45x100 AR8100
8
150ml
52x100 AR8150
4
250ml
62x100 AR8250
4
Tube type: Falcon with cap Shape: point
15ml
17x120
AF815
48
50ml
29x115
AF850
16
175ml
61x118
AF8175
4
15ml
17x120
Tube type: Corning with cap* Shape: point
250ml
60x172
AF8250
4
500ml
98x148 AF8500
4
Tube type: Falcon with cap Shape: square
12ml
17x100
AFS812
48
25ml
25x90
AFS825
28
30ml
25x110
AFS830
28
50ml
29x115
AFS850
20
15ml
17x120
AFS850
48
Tube type: Nalgene/Oakridge Shape: round
10ml
16x80
ANO810
48
30ml
26x95
ANO830
28
50ml
29x107
ANO850
24
100ml
38x106
ANO885
12
Tube type: Nalgene/Oakridge* Shape: flat
250ml
62x130
ANO8250
4
750ml
98x153
See K243/R
Tube type: Monovette Shape: square
1.1-1.4ml
8x82
AM8014
96
2.7-3ml
11x82
AM803
76
2.6-2.9ml 13x81
AM829
76
4.5-5ml
11x108
AM803
76
7.5-8.2ml 13x106
AM879
76
4.5-5ml
15x92
AM850
64
9-10ml
16x108
AM890
64
Tube type: Vacutainer Shape: round
1.6-5ml
13x75
AV816
76
4-7ml
13x100
AV850
76
8.5-10ml
16x100
AV880
64
Note: - We can not be held responsible if the tube manufacturers change the design and/or size of the tube.*
Limited capacity with sealed bucket lids (B5419 and B5319)
Rotor Fitment and Details
29
Rotor fitment
BRK5540
3
2
1
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
Place the M6
3 rotor fixing screw
(Item 2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(item 3). Ensure
all two buckets
are fully inserted
before use.
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
30
www.centurionscientificglobal.com
Microtitor Plate Rotor 4 x Standard or 2 x High Plates
Rotor
BRK5540
Buckets
Complete with buckets
Sealed Lids
Available with
Rotor type
4 x STD Plates
Tube size max
85mm x 128mm
Minimum Speed Rpm
500 Rpm
Maximum Speed Rpm
3500 Rpm
Maximum Rcf (G)
2000
Radius max cms
14.6
Sample tube angle (°)
0°C (10)
Acceleration time (secs)
30
Deceleration time (secs)
30
Autoclavable (frequency)
121°C (20)
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
31
Rotor fitment
BRK5206. BRK5212
2
3
1
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
Place the rotor
3 fixing screw
(Item2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(Item 3).
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
32
www.centurionscientificglobal.com
Rotor Details
Rotors 10,000 Rpm
Rotor
BRK5212
BRK5206
Rotor type
12 x 15ml
6 x 50ml
Tube size max
17 x 120mm
30 x 120mm
Minimum speed Rpm
500
500
Maximum speed Rpm
10,000
10,000
Maximum Rcf (G)
10,600
10,600
Radius max cms
9.5
9.5
Sample tube angle °
30
30
Acceleration time (secs)
45
45
Deceleration time (secs)
45
45
Autoclavable (frequency)
121°C (20)
121°C (20)
Refrigerated Centrifuges Only
Minimum Temperature
4°C
4°C
At maximum speed (relative to room temperature at 23°C)
Reducers
(Pack of 4)
Rotor
BRK5212
BRK5206
Part number
RM05 (5ml)
RM15(15ml)
Tube size max
13 x 80mm
17 x 120mm
Part number
RM10 (10ml)
RM25 (25ml)
Tube size max
16 x 100mm
25 x 100mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
33
Rotor fitment
BRK5210. BRK5208.
BRK5224
2
3
1
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
the rotor
3 Place
fixing screw
(Item2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(Item 3).
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
34
www.centurionscientificglobal.com
Rotor Details
Rotors 6,000 Rpm
BRK5224
Rotor
BRK5208
BRK5210
Rotor type
24 x 15ml
8 x 50ml
6 x 100ml
Size max
17 x 120mm
30 x 120mm
45 x 125mm
Minimum speed Rpm
500
500
500
Maximum speed Rpm
10,000
10,000
10,000
Maximum Rcf (G)
13,400
13,400
13,400
Radius max cms
12
12
12
Sample tube angle °
30
30
30
Acceleration time (secs)
50
50
50
Deceleration time (secs)
50
50
50
Autoclavable (frequency)
121°C (20)
121°C (20)
121°C (20)
4°C
4°C
Refrigerated Centrifuges Only
Minimum Temperature
4°C
At maximum speed (relative to room temperature at 23°C)
Reducers
Pack of 4
Pack of 6
Pack of 4
Rotor
BRK5224
BRK5208
BRK5210
Part number
RM05 (5ml)
RM15 (15ml)
RL10 (10ml)
Tube size max
13 x 80mm
17 x 120mm
16 x 100mm
Part number
RM10 (10ml)
RM25 (25ml)
RL25 (25ml)
Tube size max
13 x 100mm
25 x 100mm
25 x 100mm
RL50 (50ml)
35 x 110mm
RL85 (85ml)
39 x 110mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
35
Rotor fitment
BRK5256 (Now in silver)
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor.
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
Place the rotor
3 fixing screw
(Item2) into the
central hole of the
rotor.
4
Hand tighten with
Torque Key
(Item 3).
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
36
www.centurionscientificglobal.com
Rotor Details
Rotors 10,000 RPM
Rotor
BRK5256
Rotor type
6 x 250ml
Size max
62 x 130mm
Minimum speed Rpm
500
Maximum speed Rpm
10,000
Maximum Rcf (G)
16,100
Radius max cms
14.4
Sample tybe angle °
30
Acceleration time (secs)
60
Deceleration time (secs)
85
Autoclavable (frequecncy)
121°C (20)
Refrigerated Centrifuges Only
Minimum Temperature
4°C
At maximum speed (relative to room temperature at 23°C)
Reducers
Rotor
BRK5256
Part number
RX10 (10ml)
RX25 (25ml)
RX50 (50ml)
Tube size max
16 x 100mm
25 x 100mm
35 x 110mm
Part number
RX85 (85ml)
RX100 (100ml)
RX175 (175ml)
Tube size max
39 x 110mm
48 x 110mm
62 x 121mm
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
37
Rotor fitment
4420. 4430. 4460
1. M6 Washer
2. Rotor Screw
M6 x 20 mm
3. Torque Key
1 Place the rotor on
the shaft of the
motor
1
2
3
2 Place the M6
washer (Item 1)
over the centre
hole on the rotor.
3 Place the M6
rotor fixing screw
(Item 2) into the
central hole of the
rotor
Hand tighten with
4 Torque Key (item
3)
5
Ensure lid is fully
tightened before
use
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
38
www.centurionscientificglobal.com
Rotor Details
BRK4420 Illustrated
4420 (4 Place)
Rotor
Rotor type
4 x 0.2 to 6ml
4430 (8 Place)
4430 (12 Place)
8 x 0.2 to 6ml
12 x 0.2 to 6ml
Tube size max
Single or double
Single or double
Single or double
Minimum Speed Rpm
200
200
200
Maximum Speed Rpm
2,000
2,000
2,000
Maximum Rcf (G)
500
500
500
Radius max cms
12
12
12
Sample tube angle (°)
0
0
0
Acceleration time (secs)
25
25
25
Deceleration time (secs)
25
25
25
Autoclaveable (frequency)
121°C (10)
121°C (10)
121°C (10)
Cytology rotor includes Rotor, Sealed Lid, Clips, 24 holders with filter card & 100 slides
Accessories
4446
4444
4600
4462
Double sample holder
with card (up to 1ml)
Single sample holder
with card (up to 1ml)
Double sample holder
with card (up to 6ml)
Stainless steel clips
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
39
Rotor fitment
5450. Cell washer rotor
Please ensure that the
rotor Shaft is Clear of
any debris.
1
2 Place the rotor on the shaft.
3 Turn the knob on the rotor until
hand tight.
Make sure all the
adaptors are fully
inserted before use.
4
PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED
ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid
40
www.centurionscientificglobal.com
Rotor Details
5450. Cell washer rotor
Speed max
500-3500 Rpm
RCF max
1300 G
• High quality Polypropylene
• Autoclaveable
The rotor in this device can be
detached And efficiently function
as a holder in Which test tubes can
be fitted, Centrifuges, transported,
incubated and Emptied without the
necessity for Individual handling.
2 sets of adaptors supplied,
12 of each size
Our adaptors enable tohe rotor to
handle both 10 x 75mm and
12 x 75mm tubes.
This included rotor carries 12 tubes
but can operate with any balanced
load of fewer tubes and any
combination of sizes as long as these
are properly balanced.
This offers comlete flexibility for the
user.
PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS
Rotor Fitment and Details
41
Technical Support
Caring for your Centrifuge & Buckets
!
!
!
!
!
!
!
!
!
42
Always keep buckets & adapters clean and free from
moisture at all times, this will help prevent corrosion.
Use only official Centurion Scientific accessories with
your product Always ensure samples are distributed in a
diametric pattern to ensure the machine runs balanced.
Rotor pins should be kept clean and lubricated on a regular
basis.
It is recommended that your centrifuge is switched off and
the lid left open overnight to ensure that any moisture build
up can evaporate.
Always ensure that lids and accessories are correctly
tightened before running the unit.
Ensure that all buckets are present in the rotor when
running your centrifuge, as using the unit with an
unbalanced rotor can cause distortion and excessive wear.
Do not use any kind of scourers or abrasive material to
clean the buckets and ensure that a pH neutral cleaner is
always used.
Aluminium buckets should be kept clean and moisture free
to keep them in a good, safe operating condition.
It is recommended that regular maintenance and cleaning
of the unit and accessories is carried out.
www.centurionscientificglobal.com
Technical Support
Nomograph for RCF Determination
Example
To find the relative centrifugal
force at a radial distance of 10
cm from the center of rotation
when operating the centrifuge
at a speed of 3000 r.p.m., place
a straightedge on the chart
connecting the 10cm point on the
Rotating Radius Scale (A) with the
3000 r.p.m. point on the Speed
Scale (B).
Read the point at which the
straightedge intersects the
Relative Centrifugal Force Scale
(C) – in this case, 1000 x gravity.
Similarly, if the desired “r.c.f.” is
known, the necessary speed for
a given rotating radius may be
determined by connecting the
two known points and reading the
intersection of the straightedge
with the Speed Scale.
Equation for Calculating R.C.F.
R.C.F. = .00001118 x r x N2
R.C.F. = relative centrifugal force (gravities)
r = radius from center of rotation to bottom of tube (cm)
N = rotating speed (rev. per minute)
Rotor Fitment and Details
43
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
CHEMICAL
2-Butanol SU
S
S
2-Butyl Alcohol
S
U
S
S
2-mercaptoethanol
S
S
S
S
S
S
S
S
S
2-Propanol /
U
S
S
3-Pentanone U
M
S
Acetaldehyde
S
/
/
/
U
M
S
U
/
Acetic Acid (10%)
M
S
S
Acetic Acid (5%)
S
M
S
M
S
S
S
Acetic Acid (5%)
S
S
M
S
S
S
S
S
M
Acetic Acid (60%)
S
U
S
U
S
S
M
Acetic Acid (60%)
S
S
U
S
U
M
S
M
U
Acetic Acid (Glacial)
S
S
U
S
U
U
S
U
U
Acetic Anhydride
/
/
U
M
S
Acetone
M
S
M
S
U
S
S
M
M
Acetonitrile
S
S
S
S
U
M
S
S
S
Acetophenone /
U
M
S
Acetylene / U
M
S
Alconox™
U
U
S
S
M
S
S
S
S
Allyl Alcohol
/
/
S
/
S
S
S
/
/
Aluminium Chloride
U
U
S
S
S
S
U
Aluminium Fluoride CS
S
S
Aluminium Potassium Sulphate
C
S
S
S
Aluminium Sulphate
/
M
S
S
Aluminum Chloride
U
U
U
S
S
S
S
M
U
AmmoniaU
S
S
Ammonium Acetate
S
S
S
S
S
S
S
S
S
Ammonium Carbonate
M
S
S
S
U
S
S
S
M
Ammonium Carbonate, saturated
U
S
S
Ammonium Chloride
/
M
S
S
Ammonium Chloride, saturated
U
M
S
S
Ammonium Hydroxide (10%)
U
U
M
S
U
S
S
S
S
Ammonium Hydroxide (28%)
U
U
M
S
U
S
S
S
S
Ammonium Hydroxide (conc.)
U
U
M
/
U
S
S
S
S
Ammonium Nitrate
/
M
S
S
44
www.centurionscientificglobal.com
Ammonium Persulphate
/
M
Ammonium Phosphate
U
/
S
S
S
Ammonium Sulphate
U
M
U
S
S
Ammonium Sulphate (77%)
M
Ammonium Sulphate (sat.)
U
Ammonium Sulphide
U
Ammonium Sulphide, sat.
U
Amyl Acetate
/
U
Amyl Alcohol
S
/
S
/
S
Aniline
S
S
S
S
U
Aqua Regia
U
/
U
/
U
Barium Salts
M
U
S
S
S
Benzaldehyde /
/
U
Benzenamine /
U
Benzene
S
S
M
S
U
Benzoic Acid
/
/
S
Benzyl Alcohol
/
M
/
U
Beta-MercaptoethanolS
Boric Acid
U
S
U
S
S
Butane /U
Butyl Acetate
/
U
Calcium Bisulphate
/
U
Calcium Chloride
M
U
S
S
M
Calcium Hypochlorite
M
/
M
/
M
Calcium Salts
U
Carbon Dioxide /S
Carbon Disulphide
/
U
Carbon Tetrachloride
U
M
S
U
Cesium Acetate
M
/
S
S
S
Cesium Bromide
M
S
S
S
S
Cesium Chloride
M
S
S
S
S
Cesium Formate
M
S
S
S
S
Cesium Iodide
M
S
S
S
S
Cesium Sulphate
M
S
S
S
S
Chlorine, dry gas
/
M
Chlorine, water solution
/
/
M
S
S
S
S
S
S
S
U
M
M
U
S
M
M
U
S
U
S
S
S
S
S
S
S
S
S
M
S
S
S
S
S
S
S
U
M
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
S
S
M
S
S
U
S
S
S
S
S
/
U
/
S
S
S
/
/
/
S
S
M
S
S
S
U
U
S
S
/
S
S
S
S
S
U
S
S
S
M
S
M
U
S
S
S
S
M
S
S
M
S
S
M
S
S
M
S
S
M
S
S
M
S
S
M
S
S
Rotor Fitment and Details
45
Chlorine, wet gas
/
M
Chlorobenzene /
/
U
ChloroethaneU
Chloroform
U
U
M
S
U
Chlorosulphonic Acid
U
Chromic Acid (10%)
U
S
S
M
Chromic Acid (30%)
/
M
Chromic Acid (50%)
/
U
/
U
Chromic Acid (80%)
U
Chromic Acid -10%
U
/
U
S
M
Chromic Acid -50%
U
/
U
/
M
Citric Acid
/
M
/
M
Citric Acid (10%)
M
S
M
S
S
Copper SaltsS
Copper Sulphate
/
/
S
Cresol Mixture
S
S
S
S
U
Culture MediaS
Cyclohexane
S
S
S
S
U
Cyclohexanone /
/
U
Decalin /
U
Deoxycholate
S
S
S
S
S
Dextran
M
S
S
S
S
Dextran (Sulphate)S
Diacetone /
U
Dibutyl Phthalate
S
/
U
Dichloroethylene /U
Diethyl Ether
S
S
S
S
U
Diethyl Ketone
S
/
M
S
U
Diethylpyro-carbonate
S
S
S
S
U
Dimethylsulfoxide
S
S
S
S
U
Dioxane
M
S
M
S
U
Distilled Water
S
S
S
S
S
Ethanol (10%)
M
Ethanol (100%)
U
Ethanol (5%)
M
Ethanol (50%)
S
U
46
www.centurionscientificglobal.com
M
U
U
M
U
U
S
S
U
S
S
S
S
S
S
U
S
U
M
U
S
S
S
S
M
U
U
M
S
S
M
S
S
S
S
S
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
S
S
S
U
U
S
S
S
S
/
S
S
M
U
S
/
U
S
/
S
S
S
S
S
S
S
S
S
S
U
M
S
S
S
S
S
S
S
M
S
S
S
S
S
S
S
S
/
/
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
Ethanol (70%)
U
Ethanol (95%)
S
U
Ethanol (96%)
U
Ethanolamine /U
Ethyl Acetate
M
M
M
S
U
Ethyl Alcohol (10%)
M
Ethyl Alcohol (100%)
U
Ethyl Alcohol (5%)
M
Ethyl Alcohol (50%)
S
S
M
S
U
Ethyl Alcohol (50%)
U
Ethyl Alcohol (70%)
U
Ethyl Alcohol (95%)
S
S
M
S
U
Ethyl Alcohol (95%)
/
M
/
U
Ethyl Alcohol (96%)
/
M
/
U
Ethyl Chloride
/
U
Ethylene Dichloride
S
/
S
/
U
Ethylene Glycol
S
S
S
S
U
Ethylene Oxide (100%)
/
U
Ethylene Oxide Gas
/
M
Ethylene Oxide Vapour
S
/
/
S
M
Fatty Acids, saturated
S
U
Fatty Acids, unsaturated
U
U
Ferric Chloride
U
U
M
/
/
Ferric Sulphate /S
Ferrous Chloride
/
M
Ferrous Sulphate /S
Ficoll-Hypaque™
M
S
S
S
S
Fluoboric Acid
/
/
Formaldehyde (40%)
/
S
/
M
Formaldehyde (5%)
M
Formaldehyde -40%
M
M
S
S
S
Formalin M
Formalin (40%)
/
/
M
Formalin (5%)
M
Formic Acid
/
/
U
Formic Acid (10%)
/
M
S
S
S
S
S
S
S
S
S
S
S
S
S
S
U
U
S
U
S
S
S
S
S
S
S
S
S
S
S
S
S
M
M
S
S
S
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
S
S
S
S
M
M
S
S
S
S
S
M
S
S
S
S
U
S
/
S
/
S
S
U
/
S
S
M
S
S
S
U
S
S
S
/
M
U
S
S
S
S
S
M
S
S
/
S
S
S
M
S
S
S
S
S
Rotor Fitment and Details
47
Formic Acid (100%)
/
U
/
U
Formic Acid (3%)
/
M
Formic Acid -100%
/
S
U
/
M
Fuel Oil
/
/
M
Furfural /U
Gallic Acid
U
Gasoline /
U
Gluanidine Hydrochloric
/
S
/
S
Glutaraldehyde
S
S
S
S
S
Glycerine /
/
S
Glycerol
M
S
S
S
S
Guanidine Hydrochloride
U
U
S
S
S
Haemo-Sol™
S
S
S
S
S
Hexane
S
S
S
S
U
Hydrochloric Acid (10%)
U
M
Hydrochloric Acid (30%)
U
Hydrochloric Acid (37%)
C
U
Hydrochloric Acid (5%)
/
S
Hydrochloric Acid (50%)
/
U
/
U
Hydrochloric Acid (conc.)
U
U
U
/
U
Hydrochloric Acid -10%
U
U
U
S
U
Hydrochloric Acid -50%
U
U
U
S
U
Hydrocyanic Acid /S
Hydrofluoric Acid (10%)
/
U
/
M
Hydrofluoric Acid (100%)
U
Hydrofluoric Acid (38%)
U
Hydrofluoric Acid (50%)
/
U
/
U
Hydrofluoric Acid (60%)
U
Hydrofluoric Acid (70%)
U
Hydrofluoric Acid -10%
U
U
U
/
M
Hydrofluoric Acid -50%
U
U
U
/
U
Hydrofluoric Acid conc.
U
Hydrogen Peroxide (10%)
/
U
/
S
Hydrogen Peroxide (100%)
/
U
Hydrogen Peroxide (3%)
/
S
/
S
Hydrogen Peroxide (30%)
/
/
S
48
www.centurionscientificglobal.com
S
S
S
M
U
S
U
S
S
S
S
S
S
S
S
M
M
S
S
S
S
S
S
S
M
S
S
M
M
S
S
M
S
U
S
S
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
/
S
S
/
U
S
S
S
S
S
/
/
S
S
S
S
S
S
S
S
U
S
S
S
S
U
S
S
S
S
S
S
/
S
U
U
S
S
U
S
M
U
S
S
/
S
S
S
/
S
S
S
U
U
S
U
U
S
S
/
S
S
/
S
Hydrogen Peroxide (conc.)
U
Hydrogen Peroxide -3%
S
M
S
S
S
Hydrogen Peroxide-10%
U
U
U
S
S
Hydrogen Sulphide, dry gas
M
Hydrogen Sulphide, wet gas
C
M
Hydroquinone //
Idoacetic Acid
/
S
/
S
Iodine Solution
M
Iodoacetic Acid
S
S
S
S
S
Iso-ButanolU
Isobutyl Alcohol
/
/
S
/
S
Iso-Butyl Alcohol (10%)
U
Iso-OctaneU
Iso-Propanol (100%)
S
/
U
Isopropyl Alcohol
M
M
S
S
U
Isopropyl Alcohol (100)
U
Kerosene
S
S
S
S
M
Lacquer Thinner
U
/
U
Lactic Acid (100%)
/
/
/
/
S
Lactic Acid (20%)
/
/
/
/
S
Lactic Acid (50%)
M
Lauryl Alcohol
/
Lead Acetate
/
/
M
Magnesium Chloride
M
S
S
S
S
Magnesium Hydroxide
/
U
Magnesium Hydroxide, saturated
U
Magnesium Sulphate /S
Maleic Acid
/
M
Manganese Salts
/
Mercaptoacetic Acid
U
S
S
S
U
Mercuric Chloride
/
S
Mercury /
U
Mercury SaltsS
Methanol SU
Methanol (10%)
M
Methanol (100%)
U
U
S
S
S
S
S
S
S
S
S
S
S
U
S
S
S
M
U
S
S
S
S
S
S
S
S
S
S
/
U
S
S
S
S
S
S
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
S
S
S
S
S
M
S
S
S
S
/
S
S
M
S
S
S
S
/
S
S
S
S
S
M
S
S
U
S
S
S
M
S
S
M
S
S
S
S
S
S
M
S
S
S
S
S
S
U
S
S
S
S
S
S
S
Rotor Fitment and Details
49
Methyl Alcohol
S
S
M
S
U
Methyl Alcohol (10%)
/
M
Methyl Alcohol (100%)
/
U
Methyl Ethyl Ketone
S
S
M
S
U
Methyl Isobutyl Ketone
/
U
Methylene Chloride
U
U
S
S
U
Metrizamide™
M
S
S
S
S
Milk /S
N, N-Dimethyl- formamide
S
S
S
S
U
N/Butyl Alcohol
S
/
S
/
M
N/Butyl Phthalate
S
S
S
S
U
Naphthalene (100%)
/
U
N-Butyl Alcohol
/
S
/
U
N-Butyl Phthalate
S
S
S
Nickel ChlorideS
Nickel Salts
U
S
/
S
S
Nickel SulphateS
Nitric Acid (10%)
U
S
U
S
S
Nitric Acid (30%)
M
Nitric Acid (50%)
U
S
U
S
M
Nitric Acid (60%)
U
Nitric Acid (70%)
/
U
Nitric Acid (95%)
U
/
U
/
U
Nitric Acid conc.
/
U
Nitric Acid, fuming
U
Nitrobenzene C
/
U
N-N-Dimethylformamide
/
S
/
U
n-PropanolU
Oil (Cottons/d)
M
Oil (Lins/d)
/
U
Oil (Mineral)
/
/
U
Oil (Paraffin)
U
Oil (Silicone)
/
M
Oil (Vegetable)
M
Oils (Other)
S
/
S
S
S
Oils (Petroleum)
S
S
S
S
M
50
www.centurionscientificglobal.com
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
S
S
M
S
S
S
S
S
S
S
S
S
S
U
S
S
M
S
S
S
M
S
S
S
S
M
S
S
S
M
/
U
S
M
M
M
S
S
S
/
U
M
S
S
S
S
S
M
S
S
S
S
M
S
S
S
M
S
U
S
U
S
U
S
M
S
U
S
U
S
U
S
M
S
S
S /
S
S
S
S
M
S
U
S
U
S
M
S
S
S
S
S
/
S
U
S
U
S
Oleic Acid
S
/
U
S
S
Oxalic Acid
U
U
U
S
U
Oxalic Acid (50%)
S
OxygenS
Palmitic Acid /S
Paraffin / M
Paraffin, White
M
Perchloric Acid (10%)
/
U
/
U
Perchloric Acid (70%)
/
U
/
U
Perchloric Acid -10%
U
/
U
S
U
Perchloric Acid -70%
U
U
U
S
U
Percholric Acid (conc.)
U
Petrol /
/
U
Petroleum Ether
U
Phenol U
Phenol (5%)
U
S
M
S
U
Phenol (50%)
U
S
M
S
U
Phenol, liquefied (92%)
U
Phosphate Trichloride
/
U
Phosphoric Acid (10%)
/
U
/
S
Phosphoric Acid (25%)
M
Phosphoric Acid (30%)
U
Phosphoric Acid (50%)
U
/
U
Phosphoric Acid (85%)
/
U
Phosphoric Acid (95%)
U
Phosphoric Acid (conc.)
U
U
U
/
M
Phosphoric Acid -10%
U
U
U
S
S
Physiologic Media (Serum, Urine)
M
S
S
S
S
Picric Acid
S
S
S
S
S
Potassium AcetateS
Potassium Bromide
U
S
S
S
S
Potassium Carbonate
M
U
S
S
U
Potassium Carbonate, saturated
U
Potassium Chloride
U
S
S
S
S
Potassium Hydroxide (10%)
U
Potassium Hydroxide (45%)
U
U
S
S
S
S
S
M
M
S
M
M
M
M
U
U
M
S
M
U
U
S
S
S
S
S
S
M
S
S
S
S
S
S
S
S
S
S
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
M
U
S
S
U
S
S
S
S
S
S
/
S
/
S
U
/
S
U
U
S
S
S
S
S
U
M
S
U
U
S
S
S
/
S
S
S
S
S
S
U
M
S
U
M
S
S
S
S
U
M
S
S
S
M
S
S
S
S
S
S
U
S
S
Rotor Fitment and Details
51
Potassium Hydroxide (5%)
U
U
M
S
U
Potassium Hydroxide (conc.)
U
U
M
M
U
Potassium Iodide
/
Potassium Permanganate
S
S
S
S
S
Potassium Sulphate /S
PropaneU
Pyridine /
/
U
Pyridine (5%)
U
Pyridine (50%)
U
S
U
U
U
Rubidium Bromide
M
S
S
S
S
Rubidium Chloride
M
S
S
S
S
SeraS
Silver CyanideS
Silver Nitrate
S
/
S
Sodium Bicarbonate /S
Sodium Bisulphate /S
Sodium Borate
M
S
S
S
S
Sodium Bromide
U
S
S
S
S
Sodium Carbonate
S
/
M
Sodium Carbonate (2%)
/
S
/
M
Sodium Carbonate -2%
M
U
S
S
U
Sodium Chlorate /S
Sodium Chloride (10%)
/
S
/
S
Sodium Chloride (sat’d)
U
/
S
/
S
Sodium Chloride -10%
S
/
S
/
S
Sodium Cyanide
/
/
Sodium Dichromate
/
M
Sodium Dodecyl Sulphate
S
S
S
S
S
Sodium Hydroxide (<1%)
/
/
SS
M
Sodium Hydroxide (1%)
U
U
Sodium Hydroxide (10%)
/
U
S
M
Sodium Hydroxide (conc.)
U
Sodium Hydroxide -10%
U
/
U
M
U
Sodium Hypochlorite
M
Sodium Hypochlorite (5%)
U
U
U
S
S
52
www.centurionscientificglobal.com
S
M
S
M
S
U
M
M
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
M
M
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
M
U
U
/
U
S
S
S
M
S
S
S
S
S
S
U
S
S
M
S
S
M
S
S
S
S
S
S
S
M
S
S
M
S
S
/
S
S
S
S
S
/
/
S
S
/
S
S
S
S
S
S
S
S
/
S
S
/
S
S
M
S
S
S
M
U
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
Sodium Iodide
M
S
S
S
S
S
S
S
M
Sodium Metaborate
U
S
S
Sodium Nitrate
S
S
S
S
S
S
S
U
S
Sodium Nitrate (10%)
S
S
S
Sodium Perborate
/
/
M
S
Sodium Peroxide
/
/
S
M
S
Sodium Phosphate
M
S
S
Sodium Silicate
C
U
S
S
Sodium Sulfate
U
S
S
S
S
S
S
S
M
Sodium Sulfide
S
/
/
/
U
S
/
S
S
Sodium Sulfite
S
S
S
M
S
S
S
S
S
Sodium Sulphate
/
S
/
S
S
S
/
Sodium Sulphite
/
S
/
S
S
S
/
Sodium Thiosulphate
C
M
S
S
Solution 555 (20%)
S
S
S
S
S
S
S
S
S
Stearic Acid
S
/
S
S
S
S
S
M
M
Sucrose
M
S
S
S
S
S
S
S
S
Sucrose /
S
/ S /
Sucrose, Alkaline
M
S
S
S
U
S
S
S
M
Sulfosalicylic Acid
U
U
S
S
S
S
S
S
U
Sulfuric Acid (10%)
M
U
U
S
S
S
S
U
U
Sulfuric Acid (50%)
M
U
U
S
U
S
S
U
U
Sulfuric Acid (conc.)
M
U
U
/
U
S
S
U
U
Sulphosalicyclic Acid
/
S
/
S
S
S
/
Sulphur Chloride
/
U
U
S
Sulphur Dioxide
/
M
S
S
Sulphuric Acid (10%)
/
U
/
M
S
S
/
Sulphuric Acid (50%)
/
U
/
U
S
S
/
Sulphuric Acid (60%)
/
/
U
U
S
Sulphuric Acid (75%)
/
U
M
S
Sulphuric Acid (conc.)
/
U
/
U
S
S
/
Sulphuric Acid, fuming
U
U
S
Tannic Acid /S
S
S
Tartaric Acid
/
/
S
S
S
Tetrachloride
U
U
M
S
U
M
M
M
M
Tetrahydrofuran
S
S
U
S
U
U
S
U
S
Rotor Fitment and Details
53
TetralinU
Toluene
S
S
M
S
U
Trichloroacetic Acid
U
U
U
S
M
Trichloroacetic Acid (10%)
U
S
Trichloroethane
S
/
M
/
U
Trichloroethylene
/
/
/
/
U
Triethanolamine U
Tris Buffer (neutralpH)
U
S
S
S
S
Trisodium Phosphate
/
/
M
/
/
Triton X/100™
S
S
S
S
S
Turpentine /
/
U
Urea
S
/
S
/
M
Urine /
/
S
White Spirits
U
Xylene
S
S
M
S
U
Zephiran Chloride (1%)
S
Zephiran Chloride (7%)
M
Zinc Chloride
U
U
U
S
S
Zinc Sulphate
U
S
S
S
S
54
www.centurionscientificglobal.com
U
U
S
S
U
U
M
S
S
S
U
S
S
U
U
S
S
S
S
STAINLESS ST/L
SILICONE RUBBER
TEFLON™
POLYPROPYLENE
POLYCARBONATE
GLASS
DELRIN™
ANODIC COATING
for ALUMINIUM
Key
S - Satisfactory
M - Moderate attack
U - Unsatisfactory
ALUMINUM
Chemical resistance table
S
S
U
S
S
U
U
S
S
U
/
S
U
/
S
S
S
S
S
/
/
S
S
S
S
S
S
M
S
S
S
U
M
S
S
S
S
U
S
S
S
Head Office
The Old Stables
Church Farm
Stoughton
Chichester
West Sussex
PO18 9JL
United Kingdom.
Office 2 (Research & Service)
The Old Equine Barn, Church Farm,
Stoughton, Chichester. PO18 9JL.
Factory 1
(Ambient Manufacturing)
The Old Grain Barn, Church Farm,
Stoughton, Chichester, PO18 9JL.
Factory 2
(Refrigeration Manufacturing)
The Old Drier Barn, Church Farm,
Stoughton, Chichester, PO18 9JL.
Factory 3 (Logistics)
The Old Storage Barn, Church Farm,
Stoughton, Chichester, PO18 9JL.
Call us on: Tel +44(0)2392 631225
or email
General: info@centurionscientific.co.uk
Sales: sales@centurionscientific.co.uk
www.centurionscientificglobal.com
Follow Centurion Scientific on social media
We reserve the right to make changes with ongoing research and development
55
EC Declaration of Conformity
Declares that product: PrO-ASTM, PrO-Analytical, PrO-Research, Pro-Hospital, PrO-Ed, PrO-Vet,
PrO-Xtract, PrO-Road, PrO-PRP Centrifuge Series
Product options: Rotor plus Accessories
This instrument is fully guaranteed against manufacturing defect for a period of three years from
date of purchase.
Service and technical manual can be obtained via our website and address below.
Conform to the following Product Specifications:
IEC 1010-1, 1010-2 020, IEC 1010-2-0202 020, 7.2.101
EN 61010, EN 61010-1, 7.5 expelled parts, BS 5724, BS 7687:2.20
BS EN 61010-2-020 1195, 73/23/EEC Low voltage directive
EMC EU 89 336 (INCLUSIVE 93/68/EEC)
Supplementary Information CE Marking
The product herewith complies with the requirements of the following Directives and carries the
CE mark accordingly.
The Low voltage Directive 73.23/EEC
The EMC Directive 89/336/EEC (inclusive 93/68/EEC)
This product was tested in a typical configuration with Centurion Scientific Ltd and other test
House Facilities.
Quality Control. ISO 90001:2008
internal systems for testing sub assemblies (5) and final product.
IVD 98/79CE - European Directive.
Self certified group C.
IVD 9879EC
I hereby declare that the equipment named above has been designed to comply with the relevant
sections of the above referenced specifications. The unit complies with all the applicable essential
requirements of directives.
Certified by: K. Cooper QA Manager
Centurion
Scientific Limited
The Old Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
Certificate of Origin
Centurion Scientific Ltd of The Old Stables, Church Farm,
Stoughton, Chichester, West Sussex, PO18 9JL,
UNITED KINGDOM, hereby confirm that all centrifuges and
accessories supplied by us are of UNITED KINGDOM origin.
Signed for and on behalf of Centurion Scientific Ltd
Granted 2016
SC I ENTI FI
C
R
IO
LI
TU
MI
• CEN
CertifiCate
Of Origin
TED
• CEN
T
ON
MI
TED
RI
LI
U
Company Representative
NS
C I ENT I F
IC
Centurion
Scientific Limited
The Old Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
Centurion
Scientific Limited
Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
www.centurionscientificglobal.com
58 The Old
Rotor Fitment and Details
59
To review all of our ranges
please visit:
www.centurionscientificglobal.com
Distributed by:
Manufactured by:
Centurion Scientific Ltd
Email: info@centurionscientific.co.uk
Download