RG1100 Test Report
Date: 2020.08.21
No.: 672006007128
Test Site and Environmental conditions
Location:
The 3rd floor , No.17 Flat Gu Lou Yuan, No.618 Jinshan Road, Chang shan District, Fuzhou City
Temperature: 14.2 ℃;Humidity: 80.9
%RH;
RG1100 Main Technical Parameters
Power Supply
Two 7# dry battery
Display
128X164 LCD
Overall Dimension
70*40*95(mm)(Length * Width * height)
Instrument shell
Injection molding ABS
Sensitivity
>1.5CPS/uSv/h( Relative to 137Cs)
Temperature range
-10~+50.0℃≦±10%
Energy response
48keV~3.0MeV
Relative error
<5%( when 1mSV/h)
Measurement range
Dose Rate;0.01uSv/h~1mSv/h;
Accumulated dose ;0.00 uSv~999.9Sv
Application Field
Irradiation processing enterprises, health and epidemic prevention,
radiotherapy, nuclear laboratory and plant, import and export commodity
inspection, building materials, petrochemical industry, geological survey,
iron and steel scrap, industrial nondestructive flaw detection and related
fields.
1.Appearance inspection of product
Good packing, without damage or deformation, clean and smooth packing surface. The instrument
shell without burr, oil or oil difference.
Page 1, Total 3.
2.Software Function & Logic Testing
Testing result remark :“√”show : Qualified、“×”show : not qualified、“N”show: Non
Items
Start Test
Logic
Function
Test
Reliability
Test
Test Content
Test Standard
Test
Result
1.Start Status
Company LOGO can be showed normally
√
2.Display
Interface
Battery level, operation condition, measurement √
value are normal
3.Button
Normal without failure, without stuck key
√
1.Shutdown
Hint shutdown normally
√
2.LED Light
Set “flash” when over threshold
√
3.Buzzer
Set “noise” when over the threshold
√
4.Vibrator
Set “ Vibration” when over the threshold
√
5.Language
Select
Chinese, English selectable
√
6.Remove dose
Remove the accumulated dose
√
7.Dose Rate
Threshold
The dose rate threshold can be checked normally.
The unit can be set normally.
When alarm ,LED light flash normally,buzzer
normally,vibrator alarm normally.
√
8.Dose
Threshold
The dose threshold can be checked normally.
The unit can be set normally.
When alarm ,LED light flash normally,buzzer
normally,vibrator alarm normally.
√
9.Battery Level
The battery will from full to medium power level,
finally down to a low level and flash to alert.
√
10.Brown -out
Save
Can save the former setting or not when
accidentally brown out.
√
1. High
Tempertaure
When 48℃, operation normally
2. Low
Temperature
When -5℃, operation normally
3.EMC
Under the electromagnetic interference,
operation normally.
√
√
√
Page 2, Total 3.
Remark
3.Relative inherent error and repeatability
Ray
Relative basic error E%
Repeatability V%
γ
4.8
3.8
4.Variable Coefficient
30 minutes after starting, count it every other 1s, totally count 20, then get average P and standard
deviation, and calculate variable coefficient
5.V = 2.73 %
V =

P
=
1
P
3.92
3.76
Variable Coefficient within 5%
1 n
(Pi − P )2                              (2)

n − 1 i =1
3.65
3.85
3.72
3.68
3.74
3.78
3.78
3.82
3.69
3.88
3.67
3.94
3.83
3.91
3.91
3.82
3.81
3.74
(Unit : uSv/h)
Validity: 2020.08.21——2021.08.20
Page 3, Total 3.