Uploaded by balcknet65

41653258

advertisement
Instructions for Use
Vista 120
WARNING
To properly use this medical device, read
and comply with these Instructions for
Use.
Patient Monitor
Software 1.n
Release date April 25 2011
Screen images
Schematic renderings of screen images are used, which may differ in appearance or in
configuration from the actual screen images.
Definition of Safety Information
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING statement provides important information about a potentially
hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially hazardous
situation which, if not avoided, may result in minor or moderate injury to the user or
patient or in damage to the medical device or other property.
NOTE:
A NOTE provides additional information intended to avoid inconvenience during
operation.
Abbreviations and Symbols
For explanations refer to sections “Abbreviations” and “Symbols” in chapter
“Overview”.
Table of Contents
For Your Safety and that of Your Patients....................................................................... 1
General safety information.......................................................................................................1
Strictly follow these Instructions for Use ........................................................................1
Maintenance.......................................................................................................................1
Accessories ........................................................................................................................1
Connected devices............................................................................................................2
Not for use in areas of explosion hazard .......................................................................2
Safe connection with other electrical equipment..........................................................2
Patient safety .....................................................................................................................2
Information on Electromagnetic Compatibility ..............................................................3
Sterile accessories ............................................................................................................3
Installing accessories........................................................................................................3
Product-specific safety information.........................................................................................3
Application......................................................................................................................... 6
Intended use...............................................................................................................................6
Restrictions for use ...................................................................................................................6
Overview ............................................................................................................................ 7
Front view ...................................................................................................................................8
Side view...................................................................................................................................10
Back view.................................................................................................................................. 11
Abbreviations ...........................................................................................................................12
Symbols ....................................................................................................................................14
Vista 120 configuration ...........................................................................................................15
Getting started................................................................................................................. 16
Initial inspection .......................................................................................................................16
Installing the monitor...............................................................................................................16
Installing the monitor on a flat surface .........................................................................16
Installing the monitor on a wall ......................................................................................16
Installing the monitor on a roll stand/trolley .................................................................16
Connecting the power cord ....................................................................................................16
Checking the monitor ..............................................................................................................16
Checking the strip recorder....................................................................................................17
Setting the date and time .......................................................................................................17
System Configuration..................................................................................................... 18
Opening User Maintain Menu................................................................................................18
Entering Demo Mode ..............................................................................................................18
Selecting Lead Placement .....................................................................................................18
I
Operation ......................................................................................................................... 19
Using keys ................................................................................................................................20
Changing monitor settings .....................................................................................................21
Adjusting screen brightness...........................................................................................21
Changing date and time .................................................................................................22
Adjusting volume .....................................................................................................................22
Adjusting key volume ......................................................................................................22
Adjusting alarm volume ..................................................................................................22
Adjusting beat volume ....................................................................................................22
Checking the monitor version ................................................................................................22
Setting languages....................................................................................................................22
Understanding screens...........................................................................................................23
Calibrating screens .................................................................................................................23
Alarms.............................................................................................................................. 24
Alarm category.........................................................................................................................24
Physiological alarms .......................................................................................................24
Technical alarms ..............................................................................................................24
Prompts.............................................................................................................................24
Alarm levels ..............................................................................................................................24
Controlling alarm .....................................................................................................................25
Switching the individual alarm off..................................................................................25
Temporary alarm mute....................................................................................................25
Alarm mute .......................................................................................................................25
Setting alarm limits ..........................................................................................................25
Latching alarms........................................................................................................................26
Testing alarms ..........................................................................................................................26
Alarm information ........................................................................................................... 27
Physiological alarm information ............................................................................................27
Technical alarm information ...................................................................................................30
Prompts.....................................................................................................................................35
Adjustable range of alarm limits ............................................................................................37
Managing patients........................................................................................................... 39
Admitting a patient...................................................................................................................39
Patient category and paced status ...............................................................................39
Quick admit...............................................................................................................................40
Editing patient information......................................................................................................40
Update a patient ......................................................................................................................40
User interface .................................................................................................................. 41
Setting interface style .............................................................................................................41
Selecting display parameters ................................................................................................41
Changing waveform position .................................................................................................41
Changing interface layout ......................................................................................................41
Viewing short trend screen ....................................................................................................41
Viewing OxyCRG screen........................................................................................................41
Viewing large font screen.......................................................................................................42
II
Monitoring ECG............................................................................................................... 43
Overview ...................................................................................................................................43
ECG safety information ..........................................................................................................43
ECG display..............................................................................................................................44
Changing the size of the ECG waveform ....................................................................44
Changing the ECG filter settings...................................................................................45
Using ECG alarms...................................................................................................................45
Selecting calculation lead.......................................................................................................45
Monitoring procedure ..............................................................................................................46
Preparation .......................................................................................................................46
Connecting ECG cables .................................................................................................46
Selecting lead type ..................................................................................................................46
Installing electrodes ................................................................................................................46
Electrode placement for 3-lead .....................................................................................47
Electrode placement for 5-lead .....................................................................................47
Recommended ECG lead placement for surgical patients .......................................49
ECG menu setup .....................................................................................................................50
Setting alarm source .......................................................................................................50
Smart lead off...................................................................................................................50
Setting beat volume ........................................................................................................50
ECG display......................................................................................................................50
Setting pace status..........................................................................................................51
ECG calibration................................................................................................................51
ECG waveform settings..................................................................................................51
ST segment monitoring ..........................................................................................................51
Setting ST analysis..........................................................................................................52
ST display .........................................................................................................................52
ST analysis alarm setting ...............................................................................................52
About ST measurement points......................................................................................52
Adjusting ST and ISO measurement points ................................................................53
Arrhythmia monitoring.............................................................................................................53
Arrhythmia analysis.........................................................................................................53
ARR ANALYSIS menu ....................................................................................................55
Monitoring RESP............................................................................................................. 56
Overview ...................................................................................................................................56
RESP safety information ........................................................................................................56
Resp display.............................................................................................................................56
Electrode placement for monitoring resp.............................................................................56
Cardiac overlay ........................................................................................................................57
Chest expansion ......................................................................................................................57
Abdominal breathing ...............................................................................................................57
Selecting resp lead..................................................................................................................57
Changing hold type .................................................................................................................58
Changing the size of the respiration waveform ..................................................................58
Using Resp alarms ..................................................................................................................58
Changing the apnea time .......................................................................................................58
III
Monitoring SpO2 .............................................................................................................. 59
Overview ...................................................................................................................................59
SpO2 safety information..........................................................................................................59
Measuring SpO2 .......................................................................................................................60
Measurement procedure ........................................................................................................60
Understanding SpO2 alarms ..................................................................................................61
Adjusting alarm limits ..............................................................................................................61
Setting SpO2 as pulse source................................................................................................61
Setting pitch tone.....................................................................................................................61
Setting sensitivity.....................................................................................................................61
Monitoring PR.................................................................................................................. 62
Overview ...................................................................................................................................62
Setting PR volume...................................................................................................................62
Using pulse alarms..................................................................................................................62
Selecting the active alarm source.........................................................................................62
Monitoring NIBP .............................................................................................................. 63
Overview ...................................................................................................................................63
NIBP safety information..........................................................................................................63
Measurement limitations ........................................................................................................64
Measurement methods...........................................................................................................64
Measurement procedures ......................................................................................................65
Operation prompts...................................................................................................................66
Correcting the measurement if limb is not at heart Level .................................................66
NIBP alarm ...............................................................................................................................66
Resetting NIBP ........................................................................................................................66
Calibrating NIBP ......................................................................................................................67
Leak test ...................................................................................................................................67
Procedure for leak testing ..............................................................................................67
Monitoring TEMP............................................................................................................. 68
Overview ...................................................................................................................................68
TEMP safety information ........................................................................................................68
TEMP monitoring setup ..........................................................................................................68
Calculating temp difference ...................................................................................................68
Monitoring IBP (optional)................................................................................................ 69
Overview ...................................................................................................................................69
IBP safety information.............................................................................................................69
Monitoring procedures ............................................................................................................69
Selecting a pressure for monitoring......................................................................................70
Zeroing the pressure transducer...........................................................................................70
Troubleshooting the pressure zeroing (taking art for example) .......................................71
IBP pressure calibration .........................................................................................................71
Troubleshooting the pressure calibration.............................................................................72
IBP alarm ..................................................................................................................................73
IV
Monitoring CO2 (optional) ............................................................................................... 74
Overview ...................................................................................................................................74
CO2 safety information............................................................................................................74
Monitoring procedures ............................................................................................................75
Zeroing the sensor ..........................................................................................................75
C5 CO2 module................................................................................................................75
Setting CO2 waveform setup .................................................................................................77
Setting CO2 corrections ..........................................................................................................77
Changing CO2 alarms.............................................................................................................77
Changing apnea alarm ...........................................................................................................78
Freeze............................................................................................................................... 79
Overview ...................................................................................................................................79
Entering/exiting freeze status ................................................................................................79
Entering freeze status.....................................................................................................79
Exiting freeze status........................................................................................................79
Reviewing frozen waveform...................................................................................................80
Review.............................................................................................................................. 81
Trend graph review .................................................................................................................81
Selecting trend graph of specific parameter................................................................81
Setting resolution.............................................................................................................81
Scrolling the screen.........................................................................................................81
Switching to the trend table............................................................................................82
Record...............................................................................................................................82
Trend table review ...................................................................................................................82
Setting resolution.............................................................................................................82
Scrolling the screen.........................................................................................................82
Switching to trend graph.................................................................................................82
Recording .........................................................................................................................83
NIBP review..............................................................................................................................83
Scrolling the screen.........................................................................................................83
Recording .........................................................................................................................83
Alarm review.............................................................................................................................83
Scrolling the screen.........................................................................................................83
Selecting alarm event of specific parameter ...............................................................83
Setting time index ............................................................................................................83
Recording .........................................................................................................................84
Arr review..................................................................................................................................84
Selecting specific arrhythmia event ..............................................................................84
Scrolling the screen.........................................................................................................84
Deleting arrhythmia events ............................................................................................84
Recording .........................................................................................................................84
Calculation and titration table........................................................................................ 85
Drug calculation .......................................................................................................................85
Calculation procedures...................................................................................................85
Calculation unit ................................................................................................................86
Titration table............................................................................................................................86
V
Strip recording ................................................................................................................ 87
General information.................................................................................................................87
Performance of the recorder..................................................................................................87
Recording type.........................................................................................................................87
Starting and stopping strip recording....................................................................................88
Recorder operations and status messages.........................................................................89
Record paper requirement .............................................................................................89
Proper operation ..............................................................................................................89
Paper out ..........................................................................................................................89
Installling paper................................................................................................................89
Removing paper jam.......................................................................................................91
Other Functions .............................................................................................................. 92
Nurse call ..................................................................................................................................92
Using battery ................................................................................................................... 93
Battery power indicator...........................................................................................................93
Battery status on the main screen ........................................................................................93
Checking battery performance ..............................................................................................93
Replacing the battery ..............................................................................................................94
Recycling the battery ..............................................................................................................95
Care and cleaning ........................................................................................................... 96
General points..........................................................................................................................96
Cleaning ....................................................................................................................................96
Cleaning the monitor.......................................................................................................96
Cleaning the accessories ...............................................................................................97
Sterilization ...............................................................................................................................98
Disinfection ...............................................................................................................................98
Maintenance .................................................................................................................... 99
Inspecting .................................................................................................................................99
Maintenance task and test schedule ....................................................................................99
Replacing fuse .......................................................................................................................100
Disposal ......................................................................................................................... 101
Service policy ................................................................................................................ 102
Accessories ................................................................................................................... 103
ECG accessories...........................................................................................................103
SpO2 accessories ..........................................................................................................104
NIBP accessories ..........................................................................................................104
Temp accessories..........................................................................................................105
IBP accessories .............................................................................................................105
CO2 accessories ............................................................................................................105
Device accessories .......................................................................................................105
Spare parts .....................................................................................................................105
VI
Technical Data ............................................................................................................... 106
Classification ..........................................................................................................................106
Device specifications ............................................................................................................106
Ambient conditions ................................................................................................................106
Leakage current.....................................................................................................................106
Power supply..........................................................................................................................107
Display.....................................................................................................................................107
Recorder .................................................................................................................................107
Review.....................................................................................................................................108
ECG .........................................................................................................................................108
3-, 5-lead monitoring .....................................................................................................108
RESP....................................................................................................................................... 111
NIBP ........................................................................................................................................ 112
SpO2 ........................................................................................................................................ 113
TEMP....................................................................................................................................... 114
IBP ........................................................................................................................................... 114
CO2 .......................................................................................................................................... 114
EMC Declaration.............................................................................................................116
General information............................................................................................................... 116
Electromagnetic emissions .................................................................................................. 116
Electromagnetic environment .............................................................................................. 116
Electromagnetic immunity .................................................................................................... 117
Recommended separation distances to portable and mobile RF telecommunication
devices ....................................................................................................................................120
Default settings ............................................................................................................. 121
Patient information default settings ....................................................................................121
Alarm default settings ...........................................................................................................121
ECG default settings .............................................................................................................121
RESP.......................................................................................................................................123
SpO2........................................................................................................................................123
PR ............................................................................................................................................123
NIBP ........................................................................................................................................124
TEMP.......................................................................................................................................124
IBP ...........................................................................................................................................125
CO2 ..........................................................................................................................................125
Passwords ..................................................................................................................... 128
VII
Instructions for Use Vista 120 Patient Monitor
For Your Safety and that of Your Patients
General safety information
The following WARNING and CAUTION statements apply to general operation of the
medical device.
WARNING and CAUTION statements specific to subsystems or particular features of the
medical device appear in the respective sections of these Instructions for Use or in the
Instructions for Use of another product being used with this device.
Strictly follow these Instructions for Use
WARNING
Any use of the medical device requires full understanding and strict observation of
all sections of these Instructions for Use. The medical device must only be used
for the purpose specified under "Intended Use". Strictly observe all WARNING and
CAUTION statements throughout these Instructions for Use and all statements on
medical device labels. Failure to observe these safety information statements
constitutes a use of the medical device that is inconsistent with its intended use.
Maintenance
WARNING
The medical device must be inspected and serviced regularly by professionals
who possess the required qualifications due to their training and their experience.
Repair of the medical device must also be performed by trained personnel with
additional product-specific DrägerService training.
Dräger recommends that a service contract is obtained with DrägerService and
that all repairs are performed by DrägerService. Dräger further recommends that
only authentic Dräger repair parts are used for maintenance.
If the above are not complied with, the correct functioning of the medical device
may be compromised.
See chapter "Maintenance".
Accessories
WARNING
Only the accessories indicated on the chapter “Accessories” have been tested and
approved for use with the medical device.
Therefore, it is strongly recommended that only these accessories are used in
conjunction with the medical device. Otherwise, the correct functioning of the
medical device may be compromised.
-1-
Instructions for Use Vista 120 Patient Monitor
Connected devices
WARNING
Risk of electric shock and of device malfunction.
Any connected devices or device combinations not complying with the
requirements mentioned in these Instructions for Use may compromise the correct
functioning of the medical device. Before operating any combination of devices,
refer to and strictly comply with the Instructions for Use for all connected devices
and device combinations.
Not for use in areas of explosion hazard
WARNING
Not for use in areas of explosion hazard
This medical device is neither approved nor certified for use in areas where
combustible or explosive gas mixtures are likely to occur.
Safe connection with other electrical equipment
CAUTION
Risk of patient injury
Electrical connections to equipment not listed in these Instructions for Use or these
Assembly Instructions must only be made when approved by each respective
manufacturer.
Patient safety
The design of the medical device, the accompanying documentation, and the labeling on
the medical device are based on the assumption that the purchase and the use of the
medical device are restricted to professionals, and that certain inherent characteristics of
the medical device are known to the user. Instructions and WARNING and CAUTION
statements are therefore largely limited to the specifics of the Dräger medical device.
These Instructions for Use do not contain references to various hazards which are
obvious to professionals who operate this medical device as well as references to the
consequences of medical device misuse, and to potentially adverse effects in patients
with different underlying diseases. Medical device modification or misuse can be
dangerous.
CAUTION
Risk of patient injury
Do not make therapeutic decisions based solely on individual measured values and
monitoring parameters.
-2-
Instructions for Use Vista 120 Patient Monitor
Information on Electromagnetic Compatibility
General information on electromagnetic compatibility (EMC) according to international
EMC standard IEC 60601-1-2:
Medical electrical equipment is subject to special precautionary measures concerning
electromagnetic compatibility (EMC) and must be installed and put into operation in
accordance with the EMC information provided in this Instructions for Use.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
WARNING
Do not connect connectors with an ESD warning symbol and do not touch
the pins of such connectors without implementing ESD protective
measures. Such protective measures may include antistatic clothing and
shoes, touching a ground stud before and during connection of the pins,
or using electrically insulating and antistatic gloves. All relevant personnel
must be instructed in these ESD protective measures.
Sterile accessories
CAUTION
Do not use sterile-packaged accessories if the packaging has been opened, is damaged
or there are other signs of non-sterility. Disposable articles must not be reprocessed and
resterilized. Reuse, reprocessing, or resterilization can lead to a failure of the medical
device and cause injury to the patient.
Installing accessories
CAUTION
Install accessories to the basic device in accordance with the Instructions for Use of the
basic device. Make sure that there is a safe connection to the basic device system.
Strictly observe Assembly Instructions and Instructions for Use.
Product-specific safety information
WARNING
Before using Vista 120, patient cables, electrodes etc. should be checked.
Replacement should be taken if there is any evident defect or signs of aging which
may impair the safety or performance.
WARNING
The power receptacle must be a three-wire grounded outlet. A hospital grade outlet
is required. Never adapt the three-prong plug from the monitor to fit a two-slot
outlet.
-3-
Instructions for Use Vista 120 Patient Monitor
WARNING
Route all cables away from patient’s throat to avoid possible strangulation.
WARNING
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monitoring may
result in a hazard to the patient. Remember that the most reliable method of patient
monitoring combines close personal surveillance with correct operation of
monitoring equipment.
WARNING
When interfacing with other equipment, a test for leakage current must be
performed by qualified hospital technical personnel before using with patients.
WARNING
During monitoring, if the power supply is off and there is no battery for standby,
the monitor will be off. After reconnecting the power supply, the user should turn
on the monitor for monitoring.
WARNING
Keep away from fire immediately when leakage or foul odor is detected.
WARNING
The device and accessories are to be disposed of according to local regulations
after their useful lives. Alternatively, they can be returned to the dealer or the
manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do
NOT dispose them together with house-hold garbage. At the end of their life hand
the batteries over to the applicable collection points for the recycling of waste
batteries. For more detailed information about recycling of this product or battery,
contact your local Dräger sales representative.
WARNING
This equipment is not intended for home usage.
WARNING
Devices connected to the equipment must meet the requirements of the applicable
IEC standards (e.g., IEC 60950 “Safety standards for information technology
equipment” and IEC 60601-1 “Safety standards for medical electrical equipment”)
The system configuration must meet the requirements of the IEC 60601-1-1
“Medical electrical systems” standard. Any personnel who connect devices to the
equipment’s signal input/output port is responsible for providing evidence that the
safety certification of the devices has been performed in accordance to the IEC
60601-1-1.
CAUTION
Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
-4-
Instructions for Use Vista 120 Patient Monitor
CAUTION
Do not use autoclave or gas to sterilize the monitor, recorder or any accessories.
CAUTION
Disposable devices are intended for single use only. They should not be reused as
performance could degrade or contamination could occur.
CAUTION
Remove a battery whose life cycle has expired from the monitor immediately.
CAUTION
Avoid liquid splash on the device. The temperature must be kept between 5 and 40 °C
while working. And it should be kept between -20 and +55°C during transportation and
storage.
NOTE
• Position the device in a location where the operator can easily see the screen and
access the operating controls.
•
The monitor can only be used on one patient at a time.
•
If the monitor gets damp, put it in dry circumstance to dry it until it can work normally.
If liquid pours on the monitor, contact DrägerService.
•
Do not use this monitor for diagnostic purposes.
•
When the monitor is shut off, only the patient information and alarm settings can be
saved.
•
The pictures and dialog boxes in these Instructions for Use are for reference only.
•
Regular preventive maintenance should be carried every two years. The user is
responsible for any requirements specific to their country.
-5-
Instructions for Use Vista 120 Patient Monitor
Application
Intended use
The Vista 120 monitors parameters such as ECG (3-lead or 5-lead selectable), respiration (RESP),
functional arterial oxygen saturation (SpO2), invasive or non-invasive blood pressure (dual-IBP,
NIBP), temperature (dual-TEMP), and expired CO2.
Vista 120 is intended to be used only under regular supervision of clinical personnel. It is
applicable to adult, pediatric, and neonatal usage in a hospital environment and during patient
transport inside a healthcare facility.
Restrictions for use
CAUTION
Device for use in health care facilities only and exclusively by persons with specific
training and experience in its use.
-6-
Instructions for Use Vista 120 Patient Monitor
Overview
Vista 120 is optimized for surgical, cardiac, medical and neonatal care environments, and can
store data for both trends and events. The user can also view and record graphical and tabular
trends (vital signs).
Vista 120 has a 15-inch TFT color flat panel display. Up to 11 waveforms can be displayed on
the screen.
-7-
Instructions for Use Vista 120 Patient Monitor
Front view
A
B
C
D
E
F
G
H
J
I
A
B
Alarm
indicator
Alarm Mute
When an alarm occurs, the alarm indicator will light or flash. The alarm
level is color coded.
Press this button to pause the alarm. All audio alarms stop and the
are displayed in the
Temporary Alarm Mute **s and the symbol
information area. If the user presses the button again or if the pause time
ends, the system will resume normal monitoring status and the Temporary
Alarm Pause **s and icon will clear.
D
E
NIBP
measurement
Trend
Freeze
F
Recording
G
H
Menu
Rotary Knob
C
The symbol
is displayed in the information area. Pressing or holding
the button again resumes the alarm.
Press to inflate the cuff and perform NIBP measurement. Press again to
stop the measurement and deflate the cuff.
Press this button to enter trend table review dialog box.
In normal mode, press this button to freeze all the waveforms on the screen.
In Freeze mode, press this button to restore the waveform refreshing.
Press this button to start a real-time recording. Press again to stop
recording.
Press to return to the main dialog box.
The user can turn the rotary knob clockwise or counter-clockwise to
-8-
Instructions for Use Vista 120 Patient Monitor
highlight the desired item. Press the rotary knob to select the item.
Mains/Battery Refer to chapter “Battery power indicator” for details.
I
indicator
On/Off
When connected to the AC power supply, press the key to turn the monitor
J
on. Press the key again to turn the monitor off.
Further Alarm Mute information can be found in the chapter “Audio Off”.
-9-
Instructions for Use Vista 120 Patient Monitor
Side view
A
B
C
D
I
E
F
G
H
J
A
B
C
D
E
F
G
H
I
J
NIBP port
ECG port
etCO2 port
SpO2 port
T1 port
T2 port
IBP 1 port
IBP 2 port
Recorder
Battery compartment door
- 10 -
Instructions for Use Vista 120 Patient Monitor
Back view
A
L
K
J
B
I
C
H G
A
B
C
D
E
F
G
H
I
J
K
L
F
E
D
Fan
Speaker
Equipotential grounding terminal. If the monitor is used with other devices, connect this
terminal to eliminate potential ground differences between devices.
Security lock. Used to prevent the power cord from detaching.
Ventilation grid
Anti-theft lock
USB interface. This port is used to connect the USB device.
RJ45 network interface
RS232 interface
VGA output
Analog output. Vista 120 outputs the waveform through the auxiliary output port.
Nurse Call port. This port may be connected to an external nurse call system, not provided
by Dräger.
- 11 -
Instructions for Use Vista 120 Patient Monitor
Abbreviations
Abbr.
Explanation
Abbr.
Explanation
AC
Alternating current
Adu
Adult
AHA
American Heart Association
Art
Arterial
aVF
Left foot augmented lead
aVL
Left arm augmented lead
aVR
Right arm augmented lead
awRR
Airway respiration rate
BP
Blood pressure
BTPS
Body temperature and pressure,
saturated
CISPR
International Special Committee
on Radio Interference
CO2
Carbon dioxide
COHb
Carboxyhemoglobin
CVP
Central venous pressure
DC
Direct current
Dia
Diastolic
ECG
Electrocardiogram
EEC
European Economic Community
EMC
Electromagnetic compatibility
EMI
Electromagnetic interference
ESU
Electrosurgical unit
et
End-tidal
etCO2
End-tidal carbon dioxide
FCC
Federal Communication
Commission
FiCO2
Fraction of inspired carbon dioxide
Hb
Hemoglobin
Hb-CO
Carboxyhemoglobin
HR
Heart rate
IBP
Invasive blood pressure
ICP
Intracranial pressure
ICU
Intensive care unit
ID
Identification
IEC
International Electrotechnical
Commission
IEEE
Institute of Electrical and
Electronic Engineers
LA
Left arm
LAP
Left arterial pressure
LCD
Liquid crystal display
LED
Light emitting diode
LL
Left leg
MAP
Mean arterial pressure
MDD
Medical Device Directive
MetHb
Methemoglobin
N/A
Not applicable
Neo
Neonate
NIBP
Non-invasive blood pressure
O2
Oxygen
oxyCRG
Oxygen cardio-respirogram
PA
Pulmonary artery
PAWP
Pulmonary artery wedge pressure
Ped
Pediatric
Pleth
Plethysmogram
PR
Pulse rate
PVC
Premature ventricular complex
R
Right
- 12 -
Instructions for Use Vista 120 Patient Monitor
Abbr.
Explanation
Abbr.
Explanation
RA
Right arm
RAP
Right arterial pressure
Resp
Respiration
RHb
Reduced hemoglobin
RL
Right leg
RR
Respiration Rate
SYS
Systolic pressure
TD
Temperature difference
USB
Universal serial bus
TEMP
Temperature
- 13 -
Instructions for Use Vista 120 Patient Monitor
Symbols
Alarm Pause
USB port
NIBP measurement
RS232 port
Trend graph
VGA output, External Monitor
Freeze
Signal output port
Recording
Signal output port
Menu
Nurse call port
Locked position
Protection class type CF
(defibrillation-protected)
Battery indicator
Defibrillation-proof type BF
applied part
Alternating current
Mains indicator
Serial number
Power supply switch
Equipotential bonding
Disposal instructions
Caution
Recycle
Network port
Part number
Manufacturer
Conformitè Europèenne Directive
93/42/EEC concerning medical
products
Date of manufacture
China RoHS (Restriction of
Hazardous Substances) label
GOST label
Consult Instructions for Use
- 14 -
Instructions for Use Vista 120 Patient Monitor
Vista 120 configuration
Size (L×W×H)
Shape
Function Configuration
408 mm × 157 mm × 316 mm
Square
ECG, RESP, SpO2, NIBP, TEMP, IBP,
CO2
- 15 -
Instructions for Use Vista 120 Patient Monitor
Getting started
Initial inspection
Before unpacking, check the packaging and check that there are no signs of mishandling or
damage. If the shipping cartons are damaged, contact your local supplier for assistance.
Open the package carefully and remove the monitor and accessories. Check that all correct
options and accessories are included.
If you have any question, contact your local supplier.
Installing the monitor
Installing the monitor on a flat surface
Place the monitor on a flat surface. Make sure the surface does not vibrate and is free of corrosive
medicine and dust.
Installing the monitor on a wall
Refer to Wall-Mounting Bracket Assembly Instruction.
Installing the monitor on a roll stand/trolley
An Assembling Instruction will be delivered with the roll stand/trolley. Refer to that Assembly
Instruction for details of installing the monitor on a roll stand/trolley.
Connecting the power cord
1. Make sure the AC power supply complies with the following specifications: 100 V to 240 V~,
50 Hz/60 Hz.
2. Connect the power cord provided with the monitor to the power input of the monitor. Connect
the other end of the power cord to a grounded 3-phase power output.
Checking the monitor
Make sure there is no damage to the measurement accessories and cables. Turn on the monitor
and verify that the monitor starts normally. Make sure all alarm lamps light up and the alarm
sound is heard when turning on the monitor.
WARNING
Do not use the device if you detect any damage or if the monitor displays error
messages. Contact your hospital technical personnel or Customer Service Center
immediately.
- 16 -
Instructions for Use Vista 120 Patient Monitor
NOTE
• Check that all monitor functions are operating correctly.
•
If rechargeable batteries were provided, charge them each time before using the
device to ensure adequate power.
•
If you power the monitor off, wait at least 60 seconds before powering it on again.
Checking the strip recorder
If your monitor is equipped with a strip recorder, verify that paper is properly installed in the strip
recorder by opening the strip recorder’s door. Refer to the chapter “Recording” for details
regarding installing paper.
Setting the date and time
To set the date and time:
1. Select Menu > System Setup > Date Time Setup.
2. Adjust the date display format as desired.
3. Set the correct time of Year, Month, Day, Hour, Min and Sec from the pop-up menu and
press Exit.
- 17 -
Instructions for Use Vista 120 Patient Monitor
System Configuration
The user can not change the system configuration of the monitor. After the monitor is installed
and checked properly, a service engineer needs to change the system configuration.
Opening User Maintain Menu
1. Select the Menu item on the main interface.
2. Select Maintenance > User Maintain
3. Type the password into the displayed dialog box and press OK to enter the User Maintain
menu.
Entering Demo Mode
WARNING
The Demo Mode is for demonstration purposes only. You must not change to
Demo Mode during patient monitoring. In Demo Mode, all stored trend information
is deleted from the monitor’s memory.
To change from operating mode to demo mode:
1. Select Menu > Common Function > Demo Mode from the pop-up dialog box.
2. Type the password into the displayed dialog box.
After entering the Demo Mode, the monitor will perform the following:
•
Stops detecting alarms and generating alarm signals.
•
Stops storing data and clears previously stored data from memory.
•
All real-time data and historical data are simulated rather than actual patient data.
To exit the Demo Mode, select Menu > Common Function > Demo Mode
Selecting Lead Placement
ECG lead names have two styles: American standard (AHA) and European standard (EURO).
To set the correct style:
1. Select User Maintain > Lead Placement
2. Select AHA or EURO from the list and press the knob to confirm the selection.
- 18 -
Instructions for Use Vista 120 Patient Monitor
Operation
Everything the user needs to operate the monitor is displayed on its screen. Nearly every element
on the screen is interactive. Screen elements include measurement numerics, waveforms, screen
keys, information fields, alarms fields and menus. The monitor can be configured in a number of
different ways. For example, the user can access an item through the on-screen setup menu, via a
hard key, or via a shortcut key. This Instructions for Use describes how to access items via an
on-screen menu.
1
2
3
4
5
6
16
7
15 14
1
2
3
4
5
6
7
8
9
12 11
13
Department
Bed number
Patient name
Patient type
Alarm status area
Alarm off
Measurement value
Menu
Scroll right to display more shortcut keys
10
11
12
13
14
15
16
- 19 -
10
9 8
Date and time
Battery status symbol
AC power supply symbol
Shortcut key area
Scroll left to display more shortcut keys
Mute key
Parameter waveform
Instructions for Use Vista 120 Patient Monitor
Using keys
The monitor has four different types of keys:
Permanent keys
A permanent key is a graphical key that is permanently located at the bottom of the main screen
allowing fast, direct access to functions.
Menu
Display the main setup menu
Mute
Permanently silences auditory alarm signals
Shortcut keys
A shortcut key is a configurable graphical key located at the bottom of the main screen. It gives
the user direct access to functions. The selection of shortcut keys available on the monitor
depends on the monitor configuration and options purchased. The following functions are
available:
Admission
Admit a patient
Trend Graph
Access the trend graph review dialog box
Trend Table
Access the trend table review dialog box
AlarmReview
Access the alarm event review dialog box
NIBP Review
Access the NIBP review dialog box
ARR Review
Access the ARR review dialog box
Standard
Access the standard dialog box
TrendScreen
Access the Trend Screen dialog box
OxyCRG
Access the OxyCRG dialog box
Large Font
Access the Large Font dialog box
ModulSwitch
Access the module switch dialog box
Key Volume
Change the key volume
- 20 -
Instructions for Use Vista 120 Patient Monitor
Brightness
Adjust the screen brightness
Touch Calib
Calibrate the touch screen
IBP Zero
Zero the IBP sensor
Alarm Setup
Access the alarm setup dialog box
Beat Volume
Change the beat volume
Menu
Enter the menu
Mute
Mute the alarm
Standby
Standby mode
Hardkeys
Hardkeys are the physical keys on the front panel of the monitor. The Vista 120 has the following
hardkeys: Mute, NIBP, Trend, Freeze, Recording, and Menu. Refer also to the chapter “Front
view”.
Pop-up keys
Pop-up keys are task-related graphical keys that appear automatically on the screen when
required. For example, the confirmation pop-up key appears only when the user need to confirm a
change.
Changing monitor settings
Adjusting screen brightness
To change the screen brightness:
1. Press the Brightness key on the screen directly or.
2. Select Menu > Common Function > Brightness, and select the appropriate setting for the
screen brightness. 10 is the brightest, 1 is the least bright.
The monitor can be configured with lower brightness in standby mode and also for transport to
conserve battery power.
- 21 -
Instructions for Use Vista 120 Patient Monitor
Changing date and time
To change the date and time, refer to chapter “Setting the date and time”.
WARNING
Changing date and time will affect the storage of trend data.
Adjusting volume
Adjusting key volume
The key volume is audible when the user selects any field on the monitor screen or when the user
turns the knob. To adjust the key volume:
1. Select the Key Volume key on the screen directly or
2. Select Menu > System Setup > Key Volume, then select the appropriate setting: Five is the
loudest and Zero is the quietest.
Adjusting alarm volume
The monitor provides five levels of alarm volume: 1, 2, 3, 4 and 5.
To change the alarm volume:
1. Select Menu > Alarm Setup > Alarm Volume, and select the desired setting from the
pop-up dialog box.
Adjusting beat volume
To change the beat volume, press the Beat Volume key on the screen directly or refer to chapter
“Adjusting beat volume”.
Checking the monitor version
To check the monitor version, select Menu > Common Function > About to check the monitor
revision menu.
Setting languages
To change the language:
1. Select Menu > Maintenance > User Maintain, then type the maintenance password into the
displayed dialog box.
2. Select the Language option on the pop-up dialog box to display the language list.
3. Select the desired language from the list. To make the change permanent, restart the monitor.
- 22 -
Instructions for Use Vista 120 Patient Monitor
Understanding screens
The monitor has a set of preconfigured screens, which are optimized for common monitoring
scenarios such as OR adult or ICU neonatal. A screen defines the overall selection, size and
position of waveforms, numerics and shortcut keys displayed when the user starts the monitor.
The user can switch between different screens during monitoring. Screens do not affect alarm
settings, patient category etc. If the user changes from a complex to a less complex screen layout,
some measurements may not be visible but are still monitored in the background. For detailed
information, refer to chapter “User Interface”.
Calibrating screens
To calibrate the screen:
1. Select the Touch Calib shortcut key on the screen directly or select Menu > Maintenance >
User Maintain, then type the maintenance password into the displayed dialog box, then
select TouchScr Calibration from the pop-up dialog box.
2. The symbol
appears on the screen.
3. Click the center of the symbol
.
4. After calibration, the message Screen Calibration completed appears on the screen. Select
Exit to finish the calibration.
- 23 -
Instructions for Use Vista 120 Patient Monitor
Alarms
The following alarm information is applicable for all measurements. Refer to chapters on specific
alarms for more information.
WARNING
A potential hazard can exist if different alarm presets are used for the same or
similar equipment in different areas, e.g., an intensive care unit or an cardiac
operating room.
Alarm category
The monitor has physiological, technical, and prompt alarms.
Physiological alarms
If one or more physiological parameters monitoring the patient exceed the predefined alarm limit
(for example: APNEA and SpO2 value exceed the alarm limit), the monitor alarms. For more
information, refer to chapter “Physiological alarm information”.
Technical alarms
If, for example, the device has a low battery or defective lead the monitor will initiate a technical
alarm. Technical alarms cannot be disabled. For more information, refer to chapter “Technical
Alarm Information”.
Prompts
The device monitors processes and other functions, such as: ARR Relearning. For more
information, refer to chapter “Prompts”.
Alarm levels
There are three alarm levels consisting of high, medium, and low.
1. High level alarms
A life threatening situation exists and emergency treatment is required.
2. Medium level alarms
The patient’s vital signs or system status is abnormal. Response is required.
3. Low level alarms
The patient’s vital signs or system status is abnormal. Attention is required.
- 24 -
Instructions for Use Vista 120 Patient Monitor
Alarm sound
The high/medium/low-level alarms have the following audible characteristics:
Alarm level
High
Medium
Low
Prompt
Mode is “DO-DO-DO------DO-DO, DO-DO-DO------DO-DO”, which
is triggered once every 5 seconds. The alarm indicator flashes in red,
with a frequency of 1.4Hz~2.8Hz.
Mode is “DO-DO-DO”, which is triggered once every 25 seconds. The
alarm indicator flashes in yellow, with a frequency of 0.4Hz~0.8Hz.
Mode is “DO-”, which is triggered once every 30 seconds.
The sound pressure range for audible alarm signals is from 45 dB to 85 dB.
Controlling alarm
Switching the individual alarm off
To switch the alarm off, select XX Setup > Alarm Setup (XX stands for parameter name) and
set the alarm from the pop-up list.
Temporary alarm mute
If the monitor is in temporary alarm mute status, the monitor will:
•
Not generate an auditory alarm signal.
•
Visually display alarm information.
The temporary alarm mute symbol
a red colored background.
is displayed and the remaining pause time is displayed on
The user can set the audio pause to 60 s, 120 s, or 180 s.
Alarm mute
To mute the alarm, select Menu > Maintenance > User Maintain > Alarm Setup, and set Mute
to On.
Press the
key on the front panel for more than three seconds or the shortcut key Mute on the
screen to mute the alarm.
Setting alarm limits
WARNING
Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
- 25 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Setting alarm limits to extreme values may cause the alarm system to become
ineffective.
To change individual measurement alarm limits:
1. Select the HR Parameter area.
2. Select HR Setup > Alarm Setup. Then set the alarm limit to the desired value from the
pop-up dialog box. See the following figure:
High
Alarm Limit
Upper arrow or
lower arrow to
increase or decrease
the alarm limit
Setting value of
high alarm limit
Setting value of
low alarm limit
Low Alarm
Limit
Latching alarms
To set the alarm latch function, select Menu > Maintenance > User Maintain > Alarm Setup
and set Alarm Latch from the pull-down list. If set to On, when an alarm occurs, the monitor
will display the alarm message of the parameter in the alarm status area. If several parameters
latch alarms, each message is displayed in two second intervals.
To deselect the alarm latch, set Alarm Latch to Off. When Alarm Latch is set to Off, the latch
function is invalid.
Testing alarms
When the user switches the monitor on, a self test is started. Verify that the alarm indicator lights
and emits a single tone. The self test verifies that the visible and audible alarm indicators are
functioning correctly. For further testing of individual measurement alarms, perform the
measurement on a subject or use a simulator. If necessary, adjust the alarm limits and check that
the system is operating correctly.
- 26 -
Instructions for Use Vista 120 Patient Monitor
Alarm information
Physiological alarm information
Message
Cause
Alarm level
ECG Too Weak
Cannot detect the signal in designated time period.
High
ST-X high
ST value is above the upper alarm limit.
(X stands for I, II, III, aVR, aVL, aVF, V)
User-selectable
ST-X low
ST value is below the lower alarm limit.
(X stands for I, II, III, aVR, aVL, aVF, V)
User-selectable
ASYSTOLE
No QRS is detected for 4 consecutive seconds
User-selectable
VFIB/VTAC
Ventricular tachycardia: The fibrillation waveform lasts
for 4 consecutive seconds; or the number of continuous
Vent beats is larger than the upper limit of cluster Vent
beats (>5). The RR interval is less than 600 ms.
User-selectable
VT>2
3< the number of cluster PVCs < 5
User-selectable
COUPLET
2 consecutive PVCs
User-selectable
BIGEMINY
Vent bigeminy
User-selectable
TRIGEMINY
Vent trigeminy
User-selectable
R ON T
A type of single PVC under the condition that HR<100,
R-R interval is less than 1/3 the average interval, followed
by a compensating pause of 1.25X the average R-R
interval (the next R wave advances onto the previous T
wave).
User-selectable
PVC
Single PVCs not belonging to the type of above
mentioned PVCs.
User-selectable
TACHY
5 consecutive QRS complex,
RR interval is less than 0.5 s.
User-selectable
BRADY
5 consecutive QRS complex,
RR interval is longer than 1.5 s.
User-selectable
MISSED
BEATS
When HR is less than 100 beats/min., no heart beat is
detected during the period 1.75 times of the average RR
interval; or
User-selectable
When HR is higher than 100beat/min, no beat is detected
within 1 second.
IRR
IRREGULAR RHYTHM: The patient has irregular heart
rate, check patient’s condition, electrodes, cables and
leads.
- 27 -
User-selectable
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm level
PNC
PACE NOT CAPTURE: After the pacemaker is paced,
QRS complex can not be detected during 300 ms.
User-selectable
PNP
PACER NOT PACED: After the QRS complex, no pace
is detected during 1.75 times of RR interval.
User-selectable
VBRADY
VENTRICULAR BRADYCARDIA: The patient has
irregular HR, and his average HR is less than 60 bpm.
Check his condition, electrodes, cables and leads.
User-selectable
VENT
VENTRICULAR RHYTHM: The patient has irregular
heart rate, check patient’s condition, electrodes, cables
and leads.
User-selectable
RESP APNEA
RESP cannot be measured within specific time interval.
High
RR High
RESP value is above upper alarm limit.
User-selectable
RR Low
RESP value is below lower alarm limit.
User-selectable
HR High
HR value is above the upper alarm limit.
User-selectable
HR Low
HR value is below the lower alarm limit.
User-selectable
SpO2 High
SpO2 value is above upper alarm limit.
User-selectable
SpO2 Low
SpO2 value is below lower alarm limit.
User-selectable
SpO2 No Pulse
The signal of the measurement site is too weak, so the
monitor can not detect the pulse signal.
High
PR High
PR value is above upper alarm limit.
User-selectable
PR Low
PR value is below lower alarm limit.
User-selectable
T1 High
Value of T1 channel is above upper alarm limit.
User-selectable
T1 low
Value of T1 channel is below lower alarm limit.
User-selectable
T2 High
Value of T2 channel is above upper alarm limit.
User-selectable
T2 Low
Value of T2 channel is below lower alarm limit.
User-selectable
TD High
Value of TD channel is above upper alarm limit.
User-selectable
TEMP High
Value of TEMP is above upper alarm limit.
User-selectable
TEMP Low
Value of TEMP is below lower alarm limit.
User-selectable
SYS High
SYS value is above upper alarm limit.
User-selectable
SYS Low
SYS value is below lower alarm limit.
User-selectable
DIA High
DIA value is above upper alarm limit.
User-selectable
DIA Low
DIA value is below lower alarm limit.
User-selectable
MAP High
MAP value is above upper alarm limit.
User-selectable
- 28 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm level
MAP Low
MAP value is below lower alarm limit.
User-selectable
Art SYS High
Art SYS value is above upper alarm limit.
User-selectable
Art SYS Low
Art SYS value is below lower alarm limit.
User-selectable
Art DIA High
Art DIA value is above upper alarm limit.
User-selectable
Art DIA Low
Art DIA value is below lower alarm limit.
User-selectable
Art MAP High
Art MAP value is above upper alarm limit.
User-selectable
Art MAP Low
Art MAP value is below lower alarm limit.
User-selectable
PA SYS High
PA SYS value is above upper alarm limit.
User-selectable
PA SYS Low
PA SYS value is below lower alarm limit.
User-selectable
PA DIA High
PA DIA value is above upper alarm limit.
User-selectable
PA DIA Low
PA DIA value is below lower alarm limit.
User-selectable
PA MAP High
PA MAP value is above upper alarm limit.
User-selectable
PA MAP Low
PA MAP value is below lower alarm limit.
User-selectable
CVP MAP High CVP MAP value is above upper alarm limit.
User-selectable
CVP MAP Low
CVP MAP value is below lower alarm limit.
User-selectable
ICP MAP High
ICP MAP value is above upper alarm limit.
User-selectable
ICP MAP Low
ICP MAP value is below lower alarm limit.
User-selectable
LAP MAP High
LAP MAP value is above upper alarm limit.
User-selectable
LAP MAP Low
LAP MAP value is below lower alarm limit.
User-selectable
RAP MAP High RAP MAP value is above upper alarm limit.
User-selectable
RAP MAP Low
RAP MAP value is below lower alarm limit.
User-selectable
P1 SYS High
P1 SYS value is above upper alarm limit.
User-selectable
P1 SYS Low
P1 SYS value is below lower alarm limit.
User-selectable
P1 DIA High
P1 DIA value is above upper alarm limit.
User-selectable
P1 DIA Low
P1 DIA value is below lower alarm limit.
User-selectable
P1 MAP High
P1 MAP value is above upper alarm limit.
User-selectable
P1 MAP Low
P1 MAP value is below lower alarm limit.
User-selectable
P2 SYS High
P2 SYS value is above upper alarm limit.
User-selectable
P2 SYS Low
P2 SYS value is below lower alarm limit.
User-selectable
P2 DIA High
P2 DIA value is above upper alarm limit.
User-selectable
P2 DIA Low
P2 DIA value is below lower alarm limit.
User-selectable
- 29 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm level
P2 MAP High
P2 MAP value is above upper alarm limit.
User-selectable
P2 MAP Low
P2 MAP value is below lower alarm limit.
User-selectable
EtCO2 High
EtCO2 value is above upper alarm limit.
User-selectable
EtCO2 Low
EtCO2 value is below lower alarm limit.
User-selectable
FiCO2 High
FiCO2 value is above alarm limits.
User-selectable
AWRR High
AwRR value is above upper alarm limit.
User-selectable
AWRR Low
AwRR value is below lower alarm limit.
User-selectable
CO2 APNEA
In a specific time interval, no RESP were detected by the
CO2 module.
High
Technical alarm information
NOTE
The lead names in the table below use the AHA (American Heart Association)
denomination. For the corresponding IEC lead names, refer to chapter “Installing
Electrodes”.
Message
Cause
Alarm Level
Action Taken
ECG Lead Off
More than one ECG
electrode is off the skin or
ECG cable is disconnected
from the monitor.
Low
Make sure that all
electrodes, lead wires
and patient cables are
properly connected.
ECG V Lead Off
ECG electrode V is off the
skin or that lead wire is
disconnected.
Low
ECG LL Lead Off
ECG electrode LL is off the
skin or that lead wire is
disconnected.
Low
ECG LA Lead Off
ECG electrode LA is off the
skin or that lead wire is
disconnected.
Low
ECG RL Lead Off
ECG electrode RL is off the
skin or that lead wire is
disconnected.
Low
ECG RA Lead Off
ECG electrode RA is off the
skin or that lead wire is
disconnected.
Low
- 30 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm Level
Action Taken
ECG Signal Exceed
ECG value is beyond
measuring range.
High
Check lead connection
and patient condition
ECG Comm Fail
ECG module failure or
communication failure
High
ECG Noise
ECG signal is too noisy.
Low
Stop monitoring ECG
and notify hospital
technical personnel or
customer support.
Check lead connection
and patient condition
ECG Signal
Overflow
The amplitude of ECG is too
tall.
Low
Modify the ECG gain.
RESP Comm Fail
RESP module failure or
communication failure
High
SpO2 Sensor Off
SpO2 sensor may be
disconnected from the
patient or the monitor.
Low
Stop monitoring RESP,
and notify hospital
technical personnel or
customer support.
Ensure the sensor is
properly connected to
the patient’s body.
SpO2 No Sensor
SpO2 sensor was not
connected well or connected
to the monitor, or the
connection is loose.
Low
Ensure the monitor and
sensor are properly
connected. Reconnect
the sensor.
SpO2 Low Perfusion
The pulse signal is too weak
or the perfusion of the
measurement site is too low.
Low
Reconnect the SpO2
sensor and change the
measurement site. If
the problem continues,
notify hospital
technical personnel or
customer support.
SpO2 Comm Fail
SpO2 module failure or
communication failure
High
NIBP Comm Fail
NIBP module failure or
communication failure
High
Stop monitoring SpO2
and notify hospital
technical personnel or
customer support
Stop monitoring NIBP
and notify hospital
technical personnel or
customer support
- 31 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm Level
Action Taken
NIBP Loose Cuff
Cuff is no properly wrapped
or no cuff exists
Low
Rewrap the cuff
NIBP Weak Signal
Cuff is too loose or patient
pulse is too weak.
Low
Use a different method
to measure blood
pressure.
NIBP Excessive
Motion
Due to arm motion, signal
noise is too large or pulse
rate is not regular
Low
Ensure that the patient
being monitored is still.
NIBP Excessive
Pressure
Pressure has exceeded the
specified upper safety limit
Low
Measure again, if
failure persists, stop
monitoring NIBP and
notify hospital
technical personnel or
customer support.
NIBP Signal
Saturated
Signal amplitude too strong
Low
Stop the patient from
moving.
NIBP Init Pressure
High
The initial pressure is too
high during measuring
High
Measure again, if
failure persists, stop
monitoring NIBP and
notify hospital
technical personnel or
customer support.
NIBP Invalid Reset
The hardware pressure is too
high
Low
Measure again, if
failure persists, stop
monitoring NIBP and
notify hospital
technical personnel or
customer support.
NIBP Time Out
Measuring time has
exceeded 120 seconds
(adult/pediatric) or 90
seconds (neonatal).
Low
Measure again or use
other measuring
method.
NIBP Tube Leak
NIBP cuff or pump has a
leakage
Low
Check the NIBP cuff
and pump for leakages
NIBP Cuff type
Error
The cuff type used is
inconsistent with the patient
type
Low
Confirm the patient type
and change the cuff
- 32 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm Level
Action Taken
Air Pressure Error
Environment atmospheric
pressure abnormal or system
pressure abnormal
Low
Check whether the
airway is occluded or
pressure sensor works
properly in pressure
meter mode. If the
problem still exists,
contact your service
personnel.
NIBP Self Test Error Sensor or other hardware
errors
High
Contact your service
personnel.
NIBP Pneumatic
Leak
NIBP cuff or pump has a
leakge
Low
Check the NIBP cuff
and pump for leakages
NIBP System
Failure
Hardware abnormal
High
Contact your service
personnel.
NIBP Cuff Leak
Cuff, pumnp or airway has a
leakage
Low
NIBP cuff is not
properly connected, or
there is a leak in the
airway.
NIBP Leak Test
Error
Hardware abnormal
High
Check whether the
airway is occluded or
pressure sensor works
properly in pressure
meter mode. If the
problem still exists,
contact your service
personnel.
NIBP Range
Exceeded
The patient’s blood pressure
is beyond the measurement
range
Low
The patient’s blood
pressure is beyond the
measurement range
NIBP Air Leak
The cuff pressure cannot
reach the set value within 60
sec. or 20 sec. Neo mode
Low
Check the connections
and the wrapped cuff to
see whether they are all
well prepared.
- 33 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm Level
Action Taken
NIBP Pressure Low
The module is unable to
detect the SYSTOLIC.
Measure the pressure again.
Low
Inflate gain and retry
three times. Check
whether the patient has
an overly high blood
pressure or it is
interfered by
movement.
NIBP Pulse
Abnormal
Abnormal oscillometric
waveform
Low
Retry twice. Check for
hyperkinesia or
arrhythmia.
NIBP Pulse Signal
Weak
Pulse is too low to measure
Low
Check the patient’s
condition or the
wrapped cuff.
TEMP T1 Sensor
Off
Temperature cable of TEMP
channel 1 may be
disconnected.
Low
Make sure that the cable
is properly connected
TEMP T2 Sensor
Off
Temperature cable of TEMP
channel 2 may be
disconnected.
Low
Make sure that the cable
is properly connected.
Excessive T1
TEMP1 measuring value is
beyond measuring range.
High
Check sensor
connection and patient
condition
Excessive T2
TEMP2 measuring value is
beyond measuring range.
High
TEMP Comm Fail
TEMP module failure or
communication failure
High
YY Sensor Off
(YY stands for the
IBP label name)
IBP sensor falls off monitor.
Low
Check sensor
connection and patient
condition
Stop monitoring TEMP
and notify hospital
technical personnel or
customer support.
Make sure that cable is
properly connected.
YY Comm Fail
(YY stands for the
label name)
IBP module failure or
communication failure
High
CO2 Out Of Range
The CO2 concentration
exceeds the accuracy range
of the gas module
High
- 34 -
Stop monitoring IBP
and notify hospital
technical personnel or
customer support.
Reduce CO2
concentration
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Alarm Level
Action Taken
CO2 Sensor Faulty
CO2 module failure
High
Stop monitoring CO2
and notify hospital
technical personnel or
customer support.
CO2 Comm Fail
CO2 module failure or
communication failure
High
CO2 Ram Error
CO2 module failure
High
CO2 Rom Error
CO2 module failure
High
CO2 Zero Required
Zero calibration failure
Low
CO2 Check Adapter
The cannula is off or
disconnected
Low
Battery Low
Battery low
High
Change the battery or
recharge it
Recorder Out Of
Paper
Recorder out of paper
Low
Insert new paper
Recorder setup
needed
The user presses the
RECORD button when the
monitor is not installed with
a recorder.
Low
Notify hospital
technical personnel or
customer support to
install and set the
recorder.
CO2 Sensor Over
Temp
Prompts
Message
Cause
ECG Arr Learning
The QRS template building required for Arr. Analysis is in
process.
SpO2 Search Pulse
When the sensor is connected to the patient, the SpO2 is
analyzing the patient signal and searching for the pulse to
compute the saturation.
Manual Measuring
In manual measuring mode
Continual Measuring
In continuous measuring mode
Auto Measuring
In automatic measuring mode
- 35 -
Instructions for Use Vista 120 Patient Monitor
Message
Cause
Measure Abort
Measurement over
Calibrating
During calibrating
Calibrate Abort
Calibration over
Leak Testing
During pneumatic test
Leak Test Ok
NIBP module has passed leak test
Leak Test Abort
Pneumatic test over
Resetting
NIBP module in resetting
Please Start
NIBP module is in idle status
Done
NIBP measurement successfully done
Continual Measuring
NIBP module performs continuous measuring
Stat Measuring
NIBP module performs STAT measurement
Please Switch To Maintain
Mode
NIBP module is in normal mode, the user can not start leak
test and pressure calibration. Enter User Maintain > NIBP
Maintain and switch to Maintain Mode to perform leak test
or pressure calibration.
Please Switch To Normal
Mode
NIBP module is in maintain mode, the user can not start blood
pressure measurement. Enter User Maintain > NIBP
Maintain and switch to Normal Mode to perform blood
pressure measurement.
CO2 Standby
Changing from measuring to standby mode places the module
in energy-saving status.
CO2 Sensor Warm Up
The CO2 module is in warm-up state.
- 36 -
Instructions for Use Vista 120 Patient Monitor
Adjustable range of alarm limits
ECG alarm limits are listed as follows (unit bpm):
HR
Patient Type
ALM HI
ALM LO
ADU
300
15
PED
350
15
NEO
350
15
ST analysis alarm limits are listed as follows (unit mV):
ST
ALM HI
ALM LO
2.0
-2.0
PVCs alarm upper limits are listed as follows:
ALM HI
PVCs
ALM LO
10
RESP alarm limits are listed as follows (unit rpm):
Patient Type
ALM HI
ALM LO
ADU
120
6
PED
150
6
NEO
150
6
SpO2 alarm limits are listed as follows (unit %):
SpO2
ALM HI
ALM LO
100
0
PR alarm limit is listed as follows (unit bpm):
PR
ALM HI
ALM LO
300
30
NIBP alarm limits are listed as follows (unit mmHg):
Patient Type
ADU
ALM HI
ALM LO
SYS
270
40
DIA
215
10
- 37 -
Instructions for Use Vista 120 Patient Monitor
PED
NEO
MAP
235
20
SYS
200
40
DIA
150
10
MAP
165
20
SYS
135
40
DIA
100
10
MAP
110
20
TEMP alarm limits are listed as follows:
ALM HI
ALM LO
T1
50 °C (122 °F)
0 °C (32 °F)
T2
50 °C (122 °F)
0 °C (32 °F)
TD
50 °C (90 °F)
0 °C (0 °F)
IBP alarm limits are listed as follows (unit mmHg):
ALM HI
ALM LO
Art
300
0
CVP
40
-10
ICP
40
-10
LAP
40
-10
P1
300
-50
P2
300
-50
PA
120
-6
RAP
40
-10
CO2 alarm limits are listed as follows:
ALM HI
ALM LO
EtCO2
150 mmHg
0
FiCO2
50 mmHg
3 mmHg
AwRR
150 rpm
2 rpm
- 38 -
Instructions for Use Vista 120 Patient Monitor
Managing patients
Admitting a patient
The monitor displays patient physiological data and records the information in trends. This allows
the user to monitor unadmitted patients. It is recommended, however, that the user admit patients
to record their information in reports.
During admission, the patient category setting determines the algorithm that the monitor uses to
process and calculate measurements. These include safety limits that are applied for specific
measurements and alarm limit ranges.
To admit a patient:
1. Select the Admission shortcut key on the screen or
2. Select Menu > Patient Setup > New Patient, then a message is displayed to ask the user to
confirm to update patient.
3. Click No to cancel this operation; click Yes, the Patient Info window is displayed.
4. Enter the patient information:
•
Serial No: Enter the patient’s medical record number (MRN), for example 12345678.
•
Last name: Enter the patient’s last name (family name), for example Smith.
•
First name: Enter the patient’s first name, for example Joseph.
•
Gender: Choose Male or Female.
•
Type: Choose the patient type, either Adult, Pediat, or Neonat.
•
BloodType: Choose the patient type N/A, A, B, AB, or O.
•
Pace: Choose On or Off (The user must use On if the patient has a pacemaker).
•
Date of Birth: Enter the patient’s date of birth.
•
Date of Admission: Enter the patient’s date of admission.
•
Height: Enter the patient’s height.
•
Weight: Enter the patient’s weight.
•
Doctor: Enter any extra information about the patient or treatment.
5. Select Exit to save the changes and exit this screen.
Patient category and paced status
The patient category setting determines which algorithm the monitor uses for specific
measurements, the safety limits that are applied for specific measurements, and the alarm limit
range.
The paced setting determines whether the monitor shows pacemaker pulses. When Pace is set to
Off, pace pulses are filtered and do not show in the ECG waveform. Refer to chapter “Setting
Pace Status”.
- 39 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Changing the patient category may change the arrhythmia and NIBP alarm limits.
Always check alarm limits to make sure that they are appropriate for your patient.
WARNING
For paced patients, you must set Pace to On. If it is incorrectly set to Off, the
monitor could mistake a pace pulse for a QRS and fail to alarm during asystole.
Quick admit
If the user does not have the time or information to fully admit a patient, fill in the rest of the
patient information later.
The user can quickly admit a patient as follows:
1. Select Menu > Patient Setup > Quick Admit, a message is displayed requesting that the
user confirm the patient update.
2. Click No to cancel this operation; click Yes to continue and the Patient Info window is
displayed, choose Type and Pace and set them to the correct mode.
3. Select Exit to save the changes and exit this screen.
Editing patient information
To edit the patient information after a patient has been admitted, select Menu > Patient Setup >
Patient Info, and make the desired changes through the pop-up dialog box.
Update a patient
The user should always perform an update before starting monitoring for a new patient. To
update a patient:
•
Select Menu > Patient Setup > New Patient or Menu > Patient Setup > Quick Admit, a
menu is displayed.
•
If the user selects Yes, the monitor will update the patient information.
•
If the user selects No, the monitor won’t update the patient information and returns to
patient setup dialog box.
NOTE
Discharging a patient will clear all associated data in the monitor.
- 40 -
Instructions for Use Vista 120 Patient Monitor
User interface
Setting interface style
The user can set the interface style. The following options are available:
•
Waveform sweep
•
Monitored parameters
Only authorized personnel should set the interface style.
Selecting display parameters
The user can select the display parameters based on specific monitoring and measurement
requirements. To select the parameter:
1. Select Menu > System Setup > Module Switch.
2. Select the parameters from the pop-up dialog box.
3. Press Exit to exit the menu. The screen will adjust the parameters automatically.
Changing waveform position
The user can exchange the waveform positions of parameter A and parameter B as follows:
1. Select waveform A and display the setup menu of waveform A.
2. Select Change from the pop-up menu and select the desired label name of waveform B from
the pull-down list.
Changing interface layout
To change the interface layout:
1. Select Menu > Display Setting.
2. Select one dialog box from the pop-up menu.
3. The user can implement one kind of function screen based on the requirements. If the user
selects the Large Font option, there is no function screen to be selected.
Viewing short trend screen
To view the short trend screen, press the Trend Screen key on the screen directly or select
Menu > Display Setting > View Selection > TrendScreen.
Viewing OxyCRG screen
To view the oxyCRG screen, press the OxyCRG key on the screen or select
Menu > Display Setting > View Selection > OxyCRG.
- 41 -
Instructions for Use Vista 120 Patient Monitor
This is a NICU only feature. Monitoring of SpO2, HR and Resp is different for neonates than for
adults.
Viewing large font screen
To display the large font screen:
1. Select the Large Font key on the screen directly or.
2. Select Menu > Display Setting > View Selection > Large Font to select this display mode.
To view the large font dialog box of specific parameter, select the parameter pull-down dialog on
the dialog box (the red rectangle shown in the following figure).
- 42 -
Instructions for Use Vista 120 Patient Monitor
Monitoring ECG
Overview
The electrocardiogram (ECG) measures electrical activity in the heart and displays it both
numerically and as a waveform. This chapter also describes arrhythmia and ST segment
monitoring.
ECG safety information
WARNING
Do not touch the patient, table, or the monitor during defibrillation.
WARNING
Use only compatible ECG cables for monitoring.
WARNING
When connecting the cables and electrodes, make sure no conductive part is in
contact with the ground. Verify that all ECG electrodes, including the neutral
electrode, are securely attached to the patient but not a conductive part or ground.
WARNING
Check every day whether there is skin irritation resulted from the ECG electrodes.
If yes, replace electrodes every 24 hours or change their sites.
WARNING
Place the electrode carefully and ensure a good contact.
WARNING
Check if the lead connection is correct before monitoring. If you unplug the ECG
cable from the socket, the screen will display the error message “ECG LEAD OFF”
and an audible alarm sounds.
WARNING
When using the monitor with the defibrillator or other high-frequency equipment,
use defibrillator-proof ECG leads to avoid burns.
WARNING
When using electrosurgery (ES) equipment, do not place an electrode near the
grounding plate of the electrosurgery device: otherwise, more interference will be
evident on the ECG waveform.
WARNING
For paced patients, you must set Pace to On. If it is incorrectly set to Off, the
monitor could mistake a pace pulse for a QRS and fail to alarm during asystole.
WARNING
The simultaneous use of cardiac pacemaker and other patient-connected
equipment may cause a safety hazard.
- 43 -
Instructions for Use Vista 120 Patient Monitor
CAUTION
If patients may be defibrillated, be sure to use defibrillation-proof ECG cables, otherwise,
the ECG cables will be damaged.
NOTE
• Interference from a non-grounded instrument near the patient and ESU interference
can cause inaccuracy of the waveform.
•
IEC/EN60601-1-2 (protection against radiation is 3 V/m) specifies that the electrical
field density exceeding 1V/m may cause measurement error in various frequencies. It
is accordingly suggested to not use equipment generating electrical radiation near
ECG/RESP monitoring devices.
•
In the default settings of the monitor, the ECG waveforms are the first two waveforms
from the top in the waveform area.
•
For measurements in or near the heart connect the monitor to the potential
equalization system.
•
For protecting the environment, the used electrodes must be recycled or disposed of
properly.
ECG display
The figure below is for reference only.
The symbol “①” indicates the lead name of the displayed waveform; refer to chapter “Selecting
calculation lead”.
The symbol “②” indicates waveform gain; refer to chapter “Changing the size of the ECG
waveform”.
The symbol “③” indicates Filter setting; refer to chapter “Changing the ECG filter setting”.
Changing the size of the ECG waveform
If any of the displayed ECG waveforms are too small or clipped, the size can be changed. Select
ECG Waveform Setup > ECG Gain, then select an appropriate factor from the pop-up box to
adjust the ECG waveform.
X0.125 to make strength of ECG signal waveform of 1 mV become 1.25 mm;
- 44 -
Instructions for Use Vista 120 Patient Monitor
X0.25 to make strength of ECG signal waveform of 1 mV become 2.5 mm;
X0.5 to make strength of ECG signal waveform of 1 mV become 5 mm;
X1 to make strength of ECG signal waveform of 1 mV become 10 mm;
X2 to make strength of ECG signal waveform of 1 mV become 20 mm;
Auto let the monitor choose the optimal adjustment factor for all the ECG waveforms.
Changing the ECG filter settings
The ECG filter setting defines how ECG waveforms are smoothed. An abbreviation indicating
the filter type is shown under the lead label on the monitor display. Filter settings do not affect
ST measurement.
To change the filter setting in the ECG Setup menu, select Filter and then select the appropriate
setting.
•
Monitor: Use this mode under normal measurement conditions.
•
Surgery: The filter reduces interference to the signal. It should be used if the signal is
distorted by high- or low-frequency interference. High frequency interference usually results
in large amplitude spikes making the ECG signal look irregular. Low-frequency interference
usually leads to a wandering or rough baseline. In the operating room, the Filter reduces
artifacts and interference from HF surgical equipment. Under normal measurement conditions,
selecting Surgery may distort the QRS complexes too much and thus interfere with the
clinical evaluation of the ECG displayed on the monitor.
•
Diagnos: Use when diagnostic quality is required. The unfiltered ECG waveform is displayed
so that changes such as R-wave notching or discrete elevation or depression of the ST
segments are visible.
Using ECG alarms
ECG alarms can be switched on and off. Changes to high and low alarm limits can be made using
the same methods used in other measurement alarms. Refer to chapter “Alarms” for more
information. Only ECG special alarm features are described in this chapter.
Selecting calculation lead
On the Normal dialog box, the user can select either 3 LEADS or 5 LEADS for this item.
Normal QRS complex is defined as:
•
The normal QRS should be either completely above or below the baseline and it should not be
biphasic. For paced patients, the QRS complexes should be at least twice the height of pace
pulses.
•
The QRS should be tall and narrow.
•
The P-waves and the T-waves should be less than 0.2 mV.
- 45 -
Instructions for Use Vista 120 Patient Monitor
Monitoring procedure
Preparation
Skin is a poor conductor of electricity. Prepare the patient's skin to facilitate good electrode
conductivity.
•
Select sites with intact skin, without impairment of any kind.
•
Shave hair from sites, if necessary.
•
Wash sites thoroughly with soap and water. (Never use ether or pure alcohol, because this
increases skin impedance).
•
Rub the skin briskly to increase capillary blood flow in the tissues and remove skin scurf and
grease.
Connecting ECG cables
1.
Attach a clip or snap to the electrodes prior to placement.
2.
Place the electrodes on the patient. Before attaching, apply some conductive jelly on the
electrodes if the electrodes are not electrolyte self-supplied.
3.
Connect the electrode lead to the patient's cable.
Selecting lead type
To change the lead type:
1. Select the ECG parameter area, display the ECG Setup menu;
2. Set Lead Type to 3 Leads or 5 Leads based on the lead used.
Installing electrodes
NOTE
The following table gives the corresponding lead names according to IEC (used in
Europe) and the AHA (USA) respectively.
AHA (American Standard)
Electrode
Color
Labels
White
RA
IEC (Europe Standard)
Electrode Labels
Color
R
Red
LA
Black
L
Yellow
LL
Red
F
Green
RL
Green
N
Black
V
Brown
C
White
- 46 -
Instructions for Use Vista 120 Patient Monitor
Electrode placement for 3-lead
For the AHA standard for example, see the following figure:
•
RA: directly below the clavicle and near the right shoulder.
•
LA: directly below the clavicle and near the left shoulder.
•
LL: on the left hypogastrium.
Electrode Placement for 3-lead
Electrode placement for 5-lead
For the AHA standard for example, see the following figure:
•
RA: directly below the clavicle and near the right shoulder.
•
LA: directly below the clavicle and near the left shoulder.
•
RL: on the right hypogastrium.
•
LL: on the left hypogastrium.
•
V: on the chest, the position depends on the desired lead selection.
- 47 -
Instructions for Use Vista 120 Patient Monitor
Electrode Placement for 5-lead
NOTE
To ensure safety, all leads must be attached to the patient.
For 5-lead, attach the V electrode to one of the indicated positions as below:
•
V1
On the 4th intercostal space at the right sterna margin.
•
V2
On the 4th intercostal space at the left sterna margin.
•
V3
Midway between V2 and V4 electrodes.
•
V4
On the 5th intercostal space at the left clavicular line.
•
V5
On the left anterior axillary line, horizontal with V4 electrode.
•
V6
On the left middle axillary line, horizontal with V4 electrode.
•
V3R-V6R
On the right side of the chest in positions corresponding to those on the left.
•
VE
Over the xiphoid position.
•
V7
On the 5th intercostal space at the left posterior axillary line of back.
•
V7R
On the 5th intercostal space at the right posterior axillary line of back.
- 48 -
Instructions for Use Vista 120 Patient Monitor
V-Electrode Placement for 5-lead
Recommended ECG lead placement for surgical patients
WARNING
When using HF surgical equipment, leads should be placed in a position in equal
distance from the cautery knife and the electrosurgery grounding plate to avoid
burning the patient. HF surgical equipment wires and ECG cable must not be
tangled up.
ECG leads are used mainly for monitoring patient vital signs. When using the patient monitor
with other HF surgical equipment, use defibrillator-proof ECG leads.
The placement of the ECG leads depends on the type of surgery. For example, during open heart
surgery the electrodes can be placed laterally on the chest or back. In the operating room, artifacts
may affect the ECG waveform due to the use of ES (electrosurgery) equipment. To help reduce
this interference, the electrodes can be placed on the right and left shoulders, the right and left
sides near the abdomen and the chest lead on the left side at mid-chest. Avoid placing the
electrodes on the upper arms, which might result in a small waveform.
NOTE
• If an ECG waveform is not accurate (with electrodes are tightly attached), change the
lead.
•
Interference from a non-grounded instrument near the patient can cause inaccurate
waveforms.
- 49 -
Instructions for Use Vista 120 Patient Monitor
ECG menu setup
Setting alarm source
To change the alarm source, select ECG Setup > Alarm Source. Select from the following
options:
HR: HR is HR/PR alarm source;
PR: PR is HR/PR alarm source;
AUTO: If the Alarm Source is set to Auto, the monitor will use the heart rate from the ECG
measurement as the alarm source whenever the ECG measurement is switched on. At least one
ECG lead can be measured. The monitor will automatically switch to Pulse for the alarm source
if:
•
a valid ECG lead can no longer be measured and
•
a Pulse source is switched on and available.
The monitor uses the pulse rate from the currently active measurement as system pulse. While
Pulse is the alarm source, all arrhythmia and ECG HR alarms are switched off. If an ECG lead
becomes available again, the monitor automatically uses HR as alarm source.
Smart lead off
In 5 LEADS mode, if CH1 and CH2 are disabled due to lead off for example, the system selects
other modes to collect an ECG waveform.
To change the smart lead off setting, select ECG Setup > Smart Leadoff. Select options from
the pop-up menu.
Setting beat volume
Beat volume is derived from HR or PR, depending on the HR alarm setting. To change the beat
volume, select ECG Setup > Beat Volume, and select the desired volume from the pop-up menu.
Six selections are available: 0, 1, 2, 3, 4, 5. 5 is maximum volume. 0 is mute.
ECG display
The ECG display varies with the Lead Type. When Lead Type is set to 3 Leads, Display can be
set to Normal, and displays one ECG waveform on the main screen.
When Lead Type is set to 5 Leads, Display can be set to Normal, Full-Scr and Half-Scr. Select
Normal to display two ECG waveforms on the main screen. Select Full-Scr to display seven
ECG waveforms. Half-Scr displays seven ECG waveforms on the screen in an area of four
waveforms.
NOTE
If 3 Leads is selected in the ECG Setup menu, only Normal can be selected for Display
in the sub-menu.
- 50 -
Instructions for Use Vista 120 Patient Monitor
Setting pace status
Set the paced status correctly when starting ECG monitoring. To change the paced status in the
Setup ECG menu, select Pace to toggle between On or Off. When Pace is set to On:
-
Pace Pulse Rejection is switched on. This means that pacemaker pulses are not counted as
extra QRS complexes.
-
Paced symbol is displayed as | on the main screen.
NOTE
• When monitoring a patient with a pacemaker, set Pace to On. If monitoring a patient
without a pacemaker, set Pace to Off.
•
If Pace is set to On, the system will not perform some types of ARR analysis.
WARNING
Some pace pulses can be difficult to reject. When this happens, the pulses are
counted as a QRS complex, and could result in an incorrect HR and failure to
detect cardiac arrest or some arrhythmias. Keep pacemaker patients under close
observation.
ECG calibration
Selecting this item from the ECG Setup menu inserts a square wave into the ECG waveform that
can be used to estimate the amplitude accuracy of that waveform. Select the item again to turn off
the square wave.
NOTE
The patient is not monitored while this square wave is displayed.
ECG waveform settings
To change the speed, select ECG Wave Setup > Sweep, then select from the pop-up list. The
bigger the value, the wider the waveform.
ST segment monitoring
The monitor performs ST segment analysis on normal and atrially paced beats and calculates ST
segment elevations and depressions. This information can be displayed in the form of ST
numerics and snippets on the monitor.
ST segment monitoring function is shut off by default. It can be switched to On when necessary.
When using the ST analysis function, the result is displayed on the main screen. Refer to the
following figure.
- 51 -
Instructions for Use Vista 120 Patient Monitor
Setting ST analysis
To change ST analysis, select ECG Setup > ST Setup > ST Analysis, then select On or Off
from the pop-up list.
ST display
The screen may look different from the illustrations.
ST
I
0.08
aVR -0.09
II 0.10
aVL 0.03
III 0.02
aVF 0.06
V
0.04
NOTE
• ST Analysis can be used only in Adu mode.
•
If ST Analysis is on, the monitor must be in Diagnos mode.
•
ECG monitoring must be in Diagnos mode.
ST analysis alarm setting
The user can select Alarm > ST Alarm Setup > ST Setup to set the upper and lower alarm limit.
ALM HI can be set to 0.2 to 2.0 mV, and ALM LO to -2.0 to +0.2 mV. ALM HI must be higher
than ALM LO.
About ST measurement points
The ST value for each beat complex is the vertical difference between the isoelectric (ISO) point
and the ST point, as shown in the diagram below. The ISO point provides the baseline, and the
ST point is at the midpoint of the ST segment. The J point is where the QRS complex changes its
slope. It is a fixed distance away from the ST point and can be useful during positioning.
DEF POINT
- 52 -
Instructions for Use Vista 120 Patient Monitor
The ST and ISO measurement points need to be adjusted when monitoring is started, and if the
patient's heart rate or ECG morphology changes significantly. Always ensure that ST
measurement points are appropriate for the patient. Abnormal QRS complexes are not considered
in ST segment analysis.
Adjusting ST and ISO measurement points
Depending on the monitor’s configuration, the ST point can be positioned as well.
These two points can be adjusted by turning the knob. When adjusting ST measurement point, the
system shows the ST measurement point window. The system displays the QRS complex
template in the window and can be adjusted using the highlight bar. The user can select ISO or
ST by switching the knob left or right to move the cursor line. When the cursor is at the desired
position, the base point or the measurement point can be selected.
Arrhythmia monitoring
Arrhythmia analysis
The arrhythmia algorithm is used to monitor clinical neonatal and adult ECG. It also detects heart
rate and ventricular rhythm changes, arrhythmia events, and generates alarm information. The
arrhythmia algorithm can monitor paced and non-paced patients. Qualified personnel can use
arrhythmia analysis to evaluate patient condition (such as heart rate, PVCs frequency, rhythm and
ectopic beat). The arrhythmia algorithm can also monitor patients and generate arrhythmia
specific alarms.
The monitor can support up to 16 different arrhythmia analyses.
ARR Types
Occurring Condition
ASYSTOLE
No QRS is detected for 4 seconds
VFIB/VTAC
Ventricular tachycardia: The fibrillation waveform lasts for 4 consecutive
seconds; or the number of continuous Vent beats is larger than the upper limit
of cluster Vent beats (>5).
The RR interval is less than 600 ms.
VT>2
3< the number of cluster PVCs < 5
COUPLET
2 consecutive PVCs
BIGEMINY
Vent bigeminy
TRIGEMINY
Vent trigeminy
R ON T
A type of single PVC when HR<100,R-R interval is less than 1/3 the average
interval, followed by a compensating pause of 1.25X the average R-R interval
(the next R wave encroaches on the previous T wave).
PVC
All other single PVCs.
- 53 -
Instructions for Use Vista 120 Patient Monitor
TACHY
5 consecutive QRS complex, RR interval is less than 0.5 s.
BRADY
5 consecutive QRS complex, RR interval is longer than 1.5 s.
MISSED
BEATS
When HR is less than 100 beats/min. no heart beat is detected during the period
1.75 times of the average RR interval; or
When HR is higher than 100 beats/min. no beat is detected within 1 second.
IRR
The patient has an irregular heart rate. Check patient condition, electrodes,
cables and leads.
PNC
After the pacemaker fires, QRS complex was not be detected for 300 ms.
PNP
After the QRS complex, no pacemaker activity is detected during 1.75 times of
RR interval.
VBRADY
The patient has an irregular HR, and his average HR is less than 60 bpm.
Check patient condition, electrodes, cables and leads.
VENT
The patient has an irregular heart rate. Check patient condition, electrodes,
cables and leads.
- 54 -
Instructions for Use Vista 120 Patient Monitor
ARR ANALYSIS menu
Switching ARR analysis on and off
To switch ARR Analysis on or off: From the ARR Analysis Setup menu, select ARR Analysis
and toggle between On and Off.
PVCs alarm
Select On in the menu to enable the prompt message when an alarm occurs. Select Off to disable
the alarm function. The symbol
appears.
WARNING
When the PVCs Alarm is set to OFF, the monitor won’t give an alarm prompt even if
an alarm occurs. In order to avoid endangering the patient’s life, the user should
use this function cautiously.
ARR relearning
Select this item to start a learning procedure. ECG ARR LEARNING is displayed on the screen.
The ECG ARR LEARNING will start automatically the following:
•
Changing lead type;
•
Connecting leads;
•
Updating the patient category;
•
Starting ARR learning manually;
•
After the ARR analysis is switched on;
•
The module is set to on;
•
Calibration mode is changed to normal measurement mode;
•
Exiting the Demo mode;
•
Exiting the standby mode;
ARR alarm
The user can switch on or off all arrhythmia alarms by selecting ECG Setup > ARR Analysis >
ARR Setup > ARR Alarm. Individual arrhythmia alarms can be turned on or off. These are:
ASYSTOLE, VFIB/VTAC, R-ON-T, VT>2, COUPLET, PVC, BIGEMINY, TRIGEMINY,
TACHY, BRADY, MISSED BEATS, IRR, PNC, PNP, VBRADY and VENT.
To switch individual alarms on or off, select ECG Setup > ARR Analysis > ARR Setup > ARR
Alarm.
- 55 -
Instructions for Use Vista 120 Patient Monitor
Monitoring RESP
Overview
The monitor measures respiration from the amount of thoracic impedance between two ECG
electrodes. The change of impedance between the two electrodes, (due to the thoracic movement),
produces a respiratory waveform on the screen.
RESP safety information
WARNING
If you do not set the detection level for the respiration correctly in manual
detection mode, it may not be possible for the monitor to detect apnea. If you set
the detection level too low, the monitor is more likely to detect cardiac overlay, and
to falsely interpret cardiac overlay as respiratory activity in the case of apnea.
WARNING
The respiration measurement does not recognize obstructive and mixed apneas - it
only initiates an alarm when a pre-adjusted time has elapsed since the last
detected breath.
WARNING
If operating under conditions according to the EMC Standard EN 60601-1-2
(Radiated Immunity 3 V/m), field strengths above 1 V/m may cause erroneous
measurements at various frequencies. Therefore it is recommended to avoid the
use of electrically radiating equipment in close proximity to the respiration
measurement unit.
NOTE
The RESP monitoring is not recommended to be used on patients who are very active, as
this can cause false alarms.
Resp display
Resp
14
30
8
Electrode placement for monitoring resp
Correct patient skin preparation for electrode placement is important for Resp measurement.
Refer to the chapter on ECG for more information.
- 56 -
Instructions for Use Vista 120 Patient Monitor
The Resp signal is always measured between two of the ECG electrodes. There are two standard
ECG leads for selection: I lead (RA and LA) and II lead (RA and LL).
Electrodes Placement for 5-lead
Cardiac overlay
Cardiac activity that affects the Resp waveform is called cardiac overlay. This occurs when Resp
electrodes detect impedance changes caused by rhythmic blood flow. Correct electrode placement
can reduce cardiac overlay.
Chest expansion
Some patients, especially neonates, expand their chests laterally. In this case, place the two
respiratory electrodes in the right midaxillary and left lateral chest areas at the maximum
expansion point. This placement optimizes respiratory waveform results.
Abdominal breathing
Some patients with restricted chest movement breathe mainly abdominally. In this case, place the
LL electrode on the left abdomen at the point of maximum abdominal expansion. This placement
optimizes respiratory waveform results.
NOTE
Place the red and green electrodes diagonally to optimize the respiration waveform. Do
not position electrodes over the liver or the heart’s ventricles. This is an especially
important safety procedure for neonates.
Selecting resp lead
To change Resp lead: From the Resp Setup menu, select Resp Lead, then choose the appropriate
lead from the pop-up list.
- 57 -
Instructions for Use Vista 120 Patient Monitor
Changing hold type
To change the calculation mode: From the Resp Setup menu, set Hold Type to Manual or Auto.
When set to AUTO mode, Hold High and Hold Low are unavailable, and the monitor calculates
the respiration rate automatically. When set to Manual mode, the broken lines in RESP area can
be adjusted by using Hold High and Hold Low.
Changing the size of the respiration waveform
Select the Resp waveform area to display the Resp waveform menu:
•
Select AMP and select the desired value. Larger values increase the waveform amplitude.
•
Select Sweep: Select an appropriate setting from the pop-up list.
Using Resp alarms
The User can turn Resp alarms on and off and set high and low alarm limits using the same
procedures as described in the chapter “Alarms”.
Changing the apnea time
The apnea alarm is a high priority red alarm. The apnea alarm delay time is the period between
no detected respiratory activity and alarm activation.
1. In the Resp Setup menu, select RESP APNEA.
2. Select the desired setting from the pop-up list
- 58 -
Instructions for Use Vista 120 Patient Monitor
Monitoring SpO2
Overview
The SpO2 value is derived from the absorption of pulse blood oxygen by red and infrared light
using a finger sensor and a SpO2 measuring unit. SpO2 Plethysmogram measurement is employed
to determine the oxygen saturation of hemoglobin in the arterial blood. If, for example, 97 % of
the hemoglobin molecules in the red blood cells of the arterial blood combine with oxygen, then
the blood has a SpO2 oxygen saturation of 97 %. The SpO2 numeric on the monitor will read
97 %. The SpO2 numeric shows the percentage of hemoglobin molecules which have combined
with oxygen molecules to form oxyhemoglobin. The SpO2/PLETH parameter can also provide a
pulse rate signal and a plethysmogram waveform.
SpO2 safety information
WARNING
If the SpO2 sensor does not work properly, reconnect the sensor or use a new one.
WARNING
Do not use sterile supplied SpO2 sensors if the packaging or the sensor is
damaged.
WARNING
Prolonged and continuous monitoring may increase the risk of unexpected change
of dermal condition such as abnormal sensitivity, rubescence, vesicle, repressive
putrescence, and so on. It is especially important to check the sensor placement of
neonatal patients and patients with poor perfusion or immature dermogram by
light collimation and proper attaching strictly according to changes of the skin.
More frequent examinations may be required for different patients.
WARNING
Tissue damage may be caused by incorrect application or prolonged measurement
duration using the sensor (more than 4 hours). Inspect the sensor periodically
according to the sensor user manual.
WARNING
Neonate SpO2 sensor can only be used when required, no more than 20 min at a
time.
WARNING
Use only Dräger permitted sensors and extension cables with the oximeter. Other
sensors or extension cables may cause improper monitor performance and/or
minor personal injury.
NOTE
• Make sure the fingernail covers the light window. The wire should be placed on the
backside of the hand.
•
The amplitude of the SpO2 waveform is not proportional to the pulse volume.
- 59 -
Instructions for Use Vista 120 Patient Monitor
•
Avoid placing the sensor on extremities with an arterial catheter, or intravascular
venous infusion line.
•
Do not use a functional tester to access the SpO2 accuracy.
•
The function of SpO2 measurement does not require calibration.
•
The materials with which the patient or any other person can come into contact must
be conforming to standard ISO10993.
Measuring SpO2
1. Select the correct patient category setting (adult/pediatric and neonatal). This is used to
optimize the calculation of the SpO2 and pulse numerics.
2. During measurement, ensure that the application site:
-
has a pulsatile flow, ideally with a good circulation perfusion.
-
has not changed in its thickness, causing an improper fit of the sensor.
Measurement procedure
1. Switch on the monitor.
2. Attach the sensor to the appropriate site of the patient finger.
3. Plug the sensor extension cable connector into the SpO2 socket on the SpO2 module.
Mounting of the Sensor
WARNING
Inspect the application site every two to three hours to ensure skin quality and
correct optical alignment. If the skin quality changes, move the sensor to another
site. Change the application site at least every four hours.
NOTE
Injected dyes such as methylene blue or intravascular dyshemoglobins such as
methemoglobin and carboxyhemoglobin may lead to inaccurate measurements.
- 60 -
Instructions for Use Vista 120 Patient Monitor
Interference can be caused by:
•
High levels of ambient light or strobe lights or flashing lights (such as fire alarm lamps).
(Hint: cover application site with opaque material.)
•
Electromagnetic interference.
•
Excessive patient movement and vibration.
Understanding SpO2 alarms
This chapter describes SpO2 specific alarms. See the chapter “Alarms” for general alarm
information. SpO2 alarms are adjustable and have high and low limits.
Adjusting alarm limits
In SpO2 Setup menu:
•
Select the SpO2 High Alarm Limit and chose the desired high alarm limit.
•
Select the SpO2 Low Alarm Limit and chose the desired low alarm limit.
WARNING
High oxygen levels may predispose a premature infant to retrolental fibroplasia. If
this is a consideration do not set the high alarm limit to 100 %, which is equivalent
to switching the alarm off.
Setting SpO2 as pulse source
1. In the PR Setup menu, select PR Source;
2. Select SpO2 from the pop-up list.
Setting pitch tone
If tone modulation is on, the PR sound lowers when the SpO2 level drops. In SpO2 Setup menu,
select pitch tone to toggle between On and Off.
Setting sensitivity
Sensitivity allows the user to set the refresh frequency. High is the highest SpO2 refresh
frequency value. To change the sensitivity:
1. Select the SpO2 Setup menu;
2. Select Sensitivity and set the desired sensitivity from the pop-up list.
- 61 -
Instructions for Use Vista 120 Patient Monitor
Monitoring PR
Overview
The pulse numeric counts the arterial pulsations that result from heart mechanical activity in beats
per minute (bpm) from the measured SpO2 signal.
Setting PR volume
Six selections are available: 0, 1, 2, 3, 4, and 5. 5 is the maximum volume. 0 is mute. PR Volume
can be changed in the PR Setup menu.
Using pulse alarms
The user can change pulse rate alarm limits in the PR Setup menu by selecting PR Alarm Limit.
Pulse alarms are only generated when the active alarm source is set to Pulse. The pulse source is
set as system pulse and pulse alarms are switched on.
Selecting the active alarm source
In most cases, the HR and Pulse numerics are identical. To avoid simultaneous alarms on HR and
Pulse, the monitor uses either ECG or Pulse as its active alarm source. To change the alarm
source, select Alarm Source in the ECG/Pulse Alarms menu, then select
•
HR: If HR should be the alarm source for HR/Pulse.
•
PR: If Pulse is selected as the active alarm source, the monitor will prompt the user to
confirm his choice. If Pulse is selected as the alarm source, all arrhythmia and ECG HR
alarms are switched off.
•
AUTO: If the Alarm Source is set to Auto, the monitor will use the heart rate from the ECG
measurement as the alarm source whenever the ECG measurement is switched on and at least
one ECG lead can be measured without a technical alarm condition. The monitor will
automatically switch to Pulse as the alarm source.
- 62 -
Instructions for Use Vista 120 Patient Monitor
Monitoring NIBP
Overview
This monitor uses the oscillometric method for measuring NIBP. It can be used for adult,
pediatric and neonatal patients.
Oscillometric devices measure the amplitude of pressure changes in the occluding cuff as the cuff
deflates from above systolic pressure. The amplitude suddenly increases as the pulse breaks
through the occlusion in the artery. As the cuff pressure decreases further, the pulsations increase
in amplitude, reach a maximum (which approximates to the mean pressure), and then diminish.
In adult and pediatric mode, the blood pressure measurements determined with this device comply
with the American National Standard for Electronic or Automated Sphygmomanometers
(ANSI/AAMI SP10-1992) in relation to mean error and standard deviation, when compared to
auscultatory measurements in a representative patient population. For the auscultatory reference,
the fifth Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements determined with this device comply with the
American National Standard for Electronic or Automated Sphygmomanometers (ANSI/AAMI
SP10-1992) in relation to mean error and standard deviation, when compared to intra-arterial
measurements in a representative patient population.
NIBP safety information
The monitor and peripheral devices are protected against high-frequency interference from
defibrillators and electrosurgical units and against 50- and 60-Hz power line interference.
WARNING
Do not measure NIBP on patients with sickle-cell disease or any condition where
skin damage has occurred or is expected.
WARNING
Use clinical judgment to decide whether to perform frequent unattended blood
pressure measurements on patients with severe blood clotting disorders because
of the risk of hematoma in the limb fitted with the cuff.
WARNING
Before starting a measurement, verify that you have selected a setting appropriate
for your patient (adult, child, or neonate). An incorrect setting may be dangerous
for children, because of higher measurement pressures.
WARNING
Do not apply the cuff to a limb that has an intravenous infusion or catheter in place.
This could cause tissue damage around the catheter when infusion is slowed or
blocked during cuff inflation.
- 63 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Make sure that the air tubing connecting the blood pressure cuff and the monitor is
neither blocked nor tangled.
NOTE
• Do not start NIBP measurements during low battery conditions (the monitor may turn
off automatically).
•
If an alarm occurs or measurement fails, discontinue the measurement.
•
If you spill liquid onto the equipment or accessories, particularly if there is a chance
that it can get inside the tubing or the measurement device, contact your service
personnel.
•
Continuous use of the automatic measuring mode for short intervals may lead to the
discomfort of the patient.
Measurement limitations
Measurements are impossible with heart rate extremes of less than 40 bpm or greater than 240
bpm, or if the patient is on a heart-lung machine.
The measurement may be inaccurate or impossible in the following situations:
•
A regular arterial pressure pulse is hard to detect.
•
Patients with cardiac arrhythmias.
•
Patients with excessive and continuous movement such as shivering or convulsions.
•
Patients with rapid blood pressure changes.
•
Patients with severe shock or hypothermia that reduces blood flow to the periphery.
•
Obese patients, where a thick layer of fat surrounding a limb dampens the oscillations coming
from the artery.
•
Patients with an edematous extremity.
Measurement methods
There are three methods of measuring NIBP:
•
Manual - measurement on demand.
•
Auto - continually repeated measurements (between 1 and 480 minute adjustable interval).
•
Continual - NIBP measurements run consecutively in five minutes, and then the monitor
switches to manual mode.
- 64 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Prolonged non-invasive blood pressure measurements in Auto mode may be
associated with purport, ischemia and neuropathy in the limb wearing the cuff.
When monitoring a patient, examine the extremities of the limb frequently for
normal color, warmth and sensitivity. If any abnormality is observed, stop the
blood pressure measurements.
Measurement procedures
1.
Connect the air hose and switch on the monitor.
2.
Attach the blood pressure cuff to the patient's arm or leg and follow the following
instructions. Ensure that the cuff is completely deflated.
Attach the appropriate size cuff to the patient (For information about cuff size selection, refer
to the chapter “NIBP accessories”), and ensure the symbol "Φ" is over the artery. Ensure that
the cuff is not wrapped too tightly around the limb. Excessive tightness can cause
discoloration and ischemia of the extremity.
NOTE
The width of the cuff should be either 40 % of the limb circumference (50 % for neonates)
or 2/3 of the upper arm length. The inflatable part of the cuff should be long enough to
encircle 50-80 % of the limb. The wrong size of cuff can cause erroneous readings. If you
are unsure what size to use, consult a physician or use a large size cuff.
Cuff Usage
3.
Connect the cuff to the air tubing.
4.
Check if the patient mode is appropriate. Access the Patient Setup menu from Menu and
select Type. Turn the knob to select the desired patient Type.
5.
Select a measurement mode in the NIBP Setup menu. Refer to chapter “Operation Prompts”
for information.
6.
Press the
button on the front panel to start a measurement.
- 65 -
Instructions for Use Vista 120 Patient Monitor
Operation prompts
1.
Manual measurement
-
Access the NIBP Setup menu and set the Measure Mode item to Manual.
Press the
-
button on the front panel to start a manual measurement.
During the idle period between AUTO measurements, press the
button on the front
panel at any time to start a manual measurement. Press the
button again to stop the
manual measurement. This allows the system to continue to take AUTO measurements
according to the selected time interval.
2.
Automatic Measurement
Access the NIBP Setup menu and set the Measure Mode item to Auto.
button on the front panel to start the AUTO measurement according to the
Press the
selected time interval.
3.
Continuous measurement
Access the NIBP Setup menu and select the Continual item to start a continuous
measurement.
The continuous measurement will last 5 minutes.
4.
Stopping continuous measurement
During continuous measurement, press the
stop.
button on the front panel at any time to
Correcting the measurement if limb is not at heart Level
The following measurement corrections should be applied if the limb is not at heart level:
Add 0.75 mmHg (0.10 kPa) for each cm higher
or
Subtract 0.75 mmHg (0.10 kPa) for each cm
lower or
Add 1.9 mmHg (0.25 kPa) for each inch higher
Subtract 1.9 mmHg (0.25 kPa) for each inch
lower
NIBP alarm
When NIBP Alarm is On, the physological alarm occurs if any measurement value of Systolic
pressure, Mean pressure, Diastolic pressure is outside the alarm limits. The user can adjust the
alarm limit by accessing NIBP Setup > Alarm Setup > Sys Alarm/Mean Alarm/Dia Alarm.
Resetting NIBP
When NIBP does not work properly and the system does not generate an error message, select
Reset in the NIBP Setup menu to activate the self-test procedure. This function restores the
system.
- 66 -
Instructions for Use Vista 120 Patient Monitor
Calibrating NIBP
NIBP is not user-calibrated. Cuff-pressure transducers must be verified and calibrated, if
necessary, at least once every two years by a qualified service professional. See the Service
Guide for details.
Leak test
This item is used for the leak test. Turn the knob to select the Leak Test item in the User
Maintain > NIBP Maintain menu to start the air leakage test. When the item is selected, it will
change to Stop. If this item is selected again, the system will stop air leakage test and the item
returns to Leak Test.
WARNING
This pneumatic test, other than being specified in the EN 1060-1 standard, is to be
used by the user to simply determine whether there are air leaks in the NIBP airway.
If, at the end of the test, the system gives the prompt that the NIBP airway has air
leaks, contact the manufacturer for repair.
Procedure for leak testing
1. Connect the cuff securely with the socket for NIBP air hole.
2. Wrap the cuff around the cylinder of an appropriate size.
3. Access User Maintain > NIBP Maintain
4. Turn the knob to the Leak Test item and press the item. Then the prompt of Leak Testing
will appear indicating that the system has started performing the leak test.
5. The system will automatically inflate the pneumatic system to about 180 mmHg.
6. After 20 seconds, the system will automatically open the deflating valve, which marks the
completion of a pneumatic measurement.
If the prompt of Leak Test OK appears, it indicates that the airway is in good condition and no
air leaks exist. However if the alarm information of NIBP Cuff Leak appears, it indicates that the
airway might have air leaks. In this case, the user should check for loose connections. After
confirming secure connections, the user should re-perform the pneumatic test. If the failure
prompt still appears, contact DrägerService.
Diagram of NIBP Air Leakage Test
- 67 -
Instructions for Use Vista 120 Patient Monitor
Monitoring TEMP
Overview
Body temperature is measured using a thermistor probe (a semiconductor whose resistance
changes with temperature) placed appropriately on the patient.
Two TEMP probes can be used simultaneously to measure two TEMP values. The monitor
automatically calculates the temperature difference. The default configuration is axilla sensor for
adult.
TEMP safety information
WARNING
Verify probe cables fault detection before using the probe to monitor the patient. If
you unplug the temperature probe cable of channel 1 from the socket, the screen
will display the error message TEMP1 SENSOR OFF and the audible alarm is
activated. It is the same for the other channel.
WARNING
Handle the TEMP probe and cable carefully. When they are not in use, you should
coil the probe and cable into a loose circle. Pulling the wire inside the cable may
mechanically damage the probe and the cable.
NOTE
Disposable TEMP probe can only be used once per patient.
TEMP monitoring setup
•
If disposable TEMP probes are used, connect the TEMP cable to the monitor and then
connect the probe to the cable. With a reusable TEMP probe the probe can be connected
directly to the monitor.
•
Apply the TEMP probes securely to the patient.
•
Switch on the monitor
It takes 2 to 3 min for the body temperature to stabilize.
Calculating temp difference
The monitor can calculate and display the difference between two temperature values by
subtracting the second value from the first. The difference is labeled TD.
- 68 -
Instructions for Use Vista 120 Patient Monitor
Monitoring IBP (optional)
Overview
IBP is measured by means of a catheter inserted into the circulatory system. A pressure
transducer connected to the catheter converts the mechanical force exerted by the blood into an
electrical signal. This is displayed graphically as pressure versus time and as numerics on the
monitor screen.
The monitor measures direct blood pressure of one selected blood vessel through two channels
and displays waveforms and numerics for the measured direct blood pressure (SYS, DIA and
MAP).
For further information on the IBP Kit, refer to the accompanying Instructions for Use of the
manufacturer.
IBP safety information
WARNING
The operator should avoid contact with the conductive parts of the device when it
is connected to the patient.
WARNING
When the monitor is used with HF surgical equipment, avoid conductive
connections of the transducer and the cables to the HF surgical equipment. This is
to protect against burns to the patient.
WARNING
Disposable IBP transducer or domes should not be reused.
WARNING
If any kind of liquid, other than solution to be infused in pressure line or transducer,
is splashed on the equipment or its accessories, or enters the transducer or the
monitor, contact your service personnel immediately.
NOTE
• Use only the pressure transducer listed in the IBP accessories
Monitoring procedures
Preparatory steps for IBP measurement:
1. Connect the pressure cable to the corresponding socket and switch on the monitor.
2. Flush the system with a normal saline solution. Ensure that the system is free of air bubbles.
3. Connect the patient’s catheter to the pressure line. Ensure there is no air present in the
catheter or pressure line.
4. Position the transducer so that it is at the same level as the patient’s heart, approximately
mid-axillary line.
- 69 -
Instructions for Use Vista 120 Patient Monitor
5. For the label name selection, refer to chapter “Selecting a Pressure for Monitoring”.
6. To zero the transducer, refer to chapter “Zeroing the Pressure Transducer”.
WARNING
If there are air bubbles in the pressure line or the transducer, flush the system with
the infused solution.
Selecting a pressure for monitoring
The user can assign a pressure label to uniquely identify each pressure. The monitor then uses
that label’s stored settings (i.e., for color, scale and alarm settings) and selects which algorithm is
used to process the pressure signal. An incorrect label can lead to incorrect pressure values. To
select the label, refer to the following table:
Label
Description
ART
Arterial blood pressure
PA
Pulmonary artery pressure
CVP
Central venous pressure
ICP
Intracranial pressure
LAP
Left atrial pressure
RAP
Right atrial pressure
P1-P2
Alternative non-specific pressure labels
Zeroing the pressure transducer
To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer in
accordance with your hospital’s policy (at least once per day).
Zeroing must be performed:
•
When a new transducer or tubing is used;
•
Every time the transducer cable is reconnected to the monitor;
•
If the monitor’s pressure readings seem to be incorrect.
When using a pressure module, the zero information is stored in the module.
To zero a pressure measurement:
1. Turn off the stopcock to the patient.
2. Vent the transducer to atmospheric pressure. This compensates for the static and atmospheric
pressure exerted on the transducer.
3. In the setup menu for pressure, select Zero <Press>.
4. When the message <Press> zero done is displayed, close the stopcock to atmospheric
pressure and open the stopcock to the patient.
- 70 -
Instructions for Use Vista 120 Patient Monitor
Troubleshooting the pressure zeroing (taking art for example)
The status message lists the probable cause of an unsuccessful calibration.
Message
Corrective Action
Art ZERO FAIL
Make sure that the transducer is not attached to the patient
Art SENSOR OFF, FAIL
Make sure that transducer is not off, and then proceed
zeroing
IN DEMO, FAIL
Make sure that the monitor is not in DEMO mode. Contact
service technician if necessary
PRESSURE OVER RANGE,
FAIL
Make sure that the stopcock is vented to atmosphere. If the
problem persists, contact service technician
PULSATILE PRESSURE
ZERO FAIL
Make sure that the transducer is vented to air, not connected
to a patient, and try again.
IBP pressure calibration
•
Mercury calibration should be performed by the hospital technical personnel either whenever
a new transducer is used, or as frequently as dictated by your hospital’s policy.
•
The purpose of the calibration is to ensure that the system gives accurate measurements.
•
Before starting a mercury calibration, a zero procedure must be performed.
If the user needs to perform this procedure himself, he will need the following equipment:
•
standard sphygmomanometer
•
3-way stopcock
•
tubing (approx. 25 cm long).
The Calibration Procedure:
1.
Close the stopcock that was open to atmospheric pressure for the zero calibration.
2.
Attach the tubing to the sphygmomanometer.
3.
Ensure that connection to patient is off.
4.
Connect the 3-way connector to the 3-way stopcock that is not connected to the patient
catheter.
5.
Open the port of the 3-way stopcock to the sphygmomanometer.
6.
Select the channel to be calibrated in the menu and select the pressure value to which the IBP
is to be adjusted.
- 71 -
Instructions for Use Vista 120 Patient Monitor
7.
Inflate to make the mercury bar rise to the setup pressure value.
8.
Adjust repeatedly until the value in the menu is equal to the pressure value shown by the
mercury calibration.
9.
Press the Start button, the device will begin calibrating.
10. Wait for the calibrated result. Corresponding measures should be taken based on the prompt
information.
11. After calibration, disassemble the blood pressure tubing and the attached 3-way valve.
4
3
2
5
1
6
1: Hydrargyrum pressure meter; 2: 3-way connector; 3: 3-way stopcock; 4: Pressure transducer; 5:
Pressure transducer interface cable; 6: Monitor
IBP Calibration
Troubleshooting the pressure calibration
The status line lists the probable cause of an unsuccessful calibration.
Message
Corrective Action
Art SENSOR OFF, FAIL
Make sure that sensor is not off, and then start the calibration.
Contact service technician if necessary.
IN DEMO, FAIL
Make sure that the monitor is not in DEMO mode. Contact
service technician if necessary.
PRESSURE OVER RANGE,
FAIL
Make sure that transducer value has been selected in IBP CAL,
and then start the calibration. Contact service technician if
necessary.
Pulsatile Pressure Calibration
Fail
Make sure that pressure value shown by hydrargyrum pressure
meter is changeless. Contact service technician if necessary.
- 72 -
Instructions for Use Vista 120 Patient Monitor
IBP alarm
When Alarm Switch is On, the physological alarm occurs if any measurement value of Systolic
pressure, Mean pressure, or Diastolic pressure is outside the alarm limits. The alarm limit can be
adjusted by accessing XX Options (XX stands for the label name) > Setup > SYS Alarm >
MEAN Alarm > DIA Alarm.
- 73 -
Instructions for Use Vista 120 Patient Monitor
Monitoring CO2 (optional)
Overview
The monitor uses a Capnostat 5 CO2 module (c5) as the means for MainStream method CO2
monitoring. MainStream measurement uses a CO2 sensor attached to an airway adapter that is
directly inserted into the patient’s breathing system.
The principle of CO2 measurement is primarily based on the fact that CO2 molecules absorb
4.3μm infrared rays. Absorption intensity is proportional to the CO2 concentration of the patient
sample.
CO2 safety information
WARNING
Do not use the device in the environment with flammable anesthetic gas.
WARNING
The device should be used by trained and qualified medical personnel.
WARNING
Nitrous oxide, elevated levels of oxygen, helium, xenon, halogenated
hydrocarbons, and barometric pressure can influence the CO2 measurement.
WARNING
The monitor will be damaged if any tube from the CO2 module is disconnected, or
the air tube/the air inlet/the air outlet are plugged by water or other materials.
WARNING
The accuracy of the CO2 measurement will be affected by the following reasons:
the airway was highly obstructed; a leaky airway connection or quick variation of
environment temperature.
WARNING
Follow precautions for electrostatic discharge (ESD) and electromagnetic
interference (EMI) to and from other equipment.
WARNING
In the presence of electromagnetic devices (e.g., electrocautery), patient
monitoring may be interrupted due to electromagnetic interference.
Electromagnetic fields up to 20V/m will not adversely affect module performance.
WARNING
Do not place the sensor cables or tubing in a way that may cause tangling or
strangulation.
WARNING
Do not store the CO2 Module at temperatures less than -40 ºC or greater than
+70 ºC. Do not operate the CO2 Module at temperatures less than 0 ºC or greater
than 40 ºC.
- 74 -
Instructions for Use Vista 120 Patient Monitor
NOTE
After the low battery alarm appears, do not start the CO2 measurement. The monitor may
turn off from lack of battery power.
Monitoring procedures
Zeroing the sensor
Zeroing must be performed according to the following procedure when using a new airway
adapter:
1. Expose the sensor to room air, but not to any source of CO2. This includes the ventilator,
patient, and operator.
2. Select CO2 Setup, and change Work Mode from Standby to Measure.
3. In the CO2 Setup menu, select Zero Calibration.
4. If the system briefly displays Zeroing, the process is successful. After the zeroing calibration
is finished, you can start CO2 Monitoring. If the system displays Breath detected or Zero
required, zeroing has failed. Zero calibration must be performed again.
C5 CO2 module
NOTE
You must perform a zero calibration as described in this procedure each time you use a
new airway adapter.
C5 CO2 module
Measurement steps
1. Attach the sensor connector to the CO2 connector on the monitor.
2. Wait two minutes, allowing the sensor to reach its operating temperature and a stable thermal
condition.
3. Choose the appropriate airway adapter and connect it to the sensor head. The airway adapter
clicks into place when seated correctly.
- 75 -
Instructions for Use Vista 120 Patient Monitor
Connecting Sensor
4. To zero the sensor, refer to zeroing the sensor;
5. Install the airway adapter at the proximal end of the circuit between the elbow and the
ventilator Y-section.
Connecting Airway Adapter
WARNING
Accuracy is affected by temperature and barometric pressure.
WARNING
Installing or removing the CO2 module when the monitor is working may make the
monitor unstable. If you do it inadvertently, turn off the module in menu
immediately. The module enters STANDBY mode if you reconnect it to monitor
which it is powered on. If the readings are inaccurate, you should do calibration.
NOTE
• Replace the airway adapter if excessive moisture or secretions are observed in the
tubing or if the CO2 waveform changes unexpectedly without a change in patient
status.
- 76 -
Instructions for Use Vista 120 Patient Monitor
•
To avoid infection, use only sterilized, disinfected, or disposable airway adapters.
•
Inspect the airway adapters prior to use. Do not use if airway adapter appears to have
been damaged or broken. Observe airway adapter color coding for patient population.
•
Periodically check the flow sensor and tubing for excessive moisture or secretion
buildup.
Removing exhaust gases from the system
WARNING
Anesthetics: when using the CO2 measurement on patients who are receiving or
have recently received anesthetics connect the outlet to a scavenging system, to
avoid exposing medical staff to anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach the tube to the
mainstream sensor at the outlet connector.
Setting CO2 waveform setup
Select the CO2 waveform area to display the CO2 waveform menu:
1. Set Mode to Curve or Filled as desired.
2. Set Sweep to an appropriate value from the pop-up list. The larger the value, the faster the
speed.
Setting CO2 corrections
Temperature, water vapor in the patient’s breath, barometric pressure, and the proportions of O2,
N2O and Helium in the mixture all influence CO2 absorption. If values seem inaccurately high or
low, check that the monitor is using the appropriate corrections. Baro Press, O2 Compens,
Balance Gas and Anes Agent can be selected.
Changing CO2 alarms
This chapter describes specific CO2 alarms. See the chapter “Alarms” for general alarm
information. To change the alarm, refer to the following steps:
1. Select the CO2 Setup menu to display that menu.
2. Select etCO2 Alarm Setup, FiCO2 Alarm Setup or AwRR Alarm Setup to adjust the
alarm limit.
- 77 -
Instructions for Use Vista 120 Patient Monitor
Changing apnea alarm
This function allows the user to set an apnea alarm that monitors a time limit if a patient ceases
breathing.
1. Select the CO2 Setup menu.
2. Select Apnea Alm.
3. Choose the apnea alarm time from the pop-up list.
WARNING
Safety and effectiveness of the respiration measurement method in the detection
of apnea, particularly the apnea of prematurity and apnea of infancy, has not been
established.
- 78 -
Instructions for Use Vista 120 Patient Monitor
Freeze
Overview
When monitoring a patient, the user can freeze waveforms and examine them. Generally, the user
should not view frozen waveforms for longer than 60 seconds. The Freeze function has the
following features:
•
Freeze status can be activated on any operating screen.
•
Upon freezing, the system exits all other operating menus. The system freezes all waveforms
in the Waveform area of the Basic Screen, the Full Lead ECG waveforms and extra
waveforms on the Full Lead ECG interface (if any). The Parameter area refreshes normally.
•
The frozen waveforms can be reviewed and recorded.
Entering/exiting freeze status
Entering freeze status
In the non-Freeze status, press the
hardkey on the monitor control panel to exit the current
menu. Freeze status is entered and the pop-up Freeze menu is displayed. In Freeze status, all
waveforms are frozen and will not be refreshed.
Exiting freeze status
Executing any of the following operations will exit Freeze status:
1. Select Exit option in/from the Freeze menu;
2. Press
hardkey on the control panel again;
3. Execute any operation that triggers the adjustment of the screen or displays a new menu.
After exiting Freeze status, the system will clear screen waveforms and resume displaying
real-time waveforms. In the Screen Refresh mode, the system will sweep the waveforms from left
to right in the Waveform Area.
Press the
hardkey on the control panel, and the Freeze menu will appear on the bottom part
of the screen. At the same time, the system freezes the waveforms.
•
REC WAVE: can be set to any waveform of 8 s, such as IBP1, CO2, PLETH etc. It can also
be set to OFF.
•
Review: Used to review frozen waveforms.
•
Exit: The system closes the Frozen menu and exits Freeze status.
NOTE
Pressing the
hardkey repeatedly over a short period of time may result in
discontinuous waveforms on the screen.
- 79 -
Instructions for Use Vista 120 Patient Monitor
Reviewing frozen waveform
By moving the waveform, the user can review 120 seconds of the waveform before it was frozen.
For a waveform of less than 60 seconds, the remaining part is displayed as a straight line. The
rotary knob on the control panel can be used to move the cursor to the Review option in the
Freeze menu. Press the knob. By turning the knob left or right, frozen waveforms on the screen
will move left or right correspondingly. There is an arrow indicating upward on the right side of
the last waveform.
- 80 -
Instructions for Use Vista 120 Patient Monitor
Review
The monitor provides 120-hour trend data of all parameters, storage of 1200 NIBP measurement
results and 60 alarm events. This chapter gives detailed instruction for review of all data.
Trend graph review
•
The latest 1-hour trend displays data taken every 1 or 5 seconds.
•
The latest 120-hour trend displays data taken every 1, 5 or 10 minutes.
To review the Trend Graph, press the Trend hardkey on the front panel or select
Menu > Review > Trend Graph.
In the trend graph, the y-axis is the measurement value and x-axis represents time.
Selecting trend graph of specific parameter
The monitor can display trend graphs of different parameters. To select a different parameter’s
trend, select Menu > Review > Trend Graph and select the desired parameter name from the
pop-up list (as shown in red text on the above figure).
Setting resolution
To set an appropriate resolution, select Menu > Review > Trend Graph. Choose Resolution to
display the list and select a resolution ranging from 1 sec, 5 sec, 1 min, 5 min and 10 min.
Scrolling the screen
Not all trend graphs are displayed on the current screen. To see other trend graphs, press the
symbols
and
displayed on the trend graph.
- 81 -
Instructions for Use Vista 120 Patient Monitor
Switching to the trend table
To switch to the trend table, select Trend Table.
Record
The monitor can print the data in the current trend window. For more information, refer to
chapter “Strip recording”.
Trend table review
To review the trend table, press the Trend Table shortcut key on the screen or select
Menu > Review > Trend Table.
Setting resolution
To set an appropriate resolution, select Menu > Review > Trend Table. Choose Resolution to
display the list and select a resolution ranging from 1 sec, 5 sec, 1 min, 5 min and 10 min, 30
min and 60 min.
Scrolling the screen
Not all trend tables can be displayed on the current screen. To see other measurement trend tables,
press the symbols
,
,
and
displayed on the trend graph.
Switching to trend graph
The user can switch to the trend graph on the Trend Table dialog box.
Select Menu > Review > Trend Graph and select the Trend Graph option from the pop-up
dialog box.
- 82 -
Instructions for Use Vista 120 Patient Monitor
Recording
The monitor can make a tabular trend recording of the data in the current trend graph window.
The report uses the current trend interval settings. For more information, refer to chapter
“Recording”.
NIBP review
To review the NIBP measurement data, select the NIBP Review key on the screen or select
Menu > Review > NIBP Review. The NIBP Review window is displayed.
Scrolling the screen
Not all measurement data is displayed on the current screen. To see other measurement data,
press the symbols
and
displayed on the NIBP Review dialog box.
Recording
The monitor can record the measurement data in the NIBP review window. For more information,
refer to chapter “Recording”.
Alarm review
The monitor can store up to 60 alarm events and displays the last 10 alarm events.
To review the alarm event, select the Alarm Review key on the screen or select Menu > Review
> Alarm Review. The Alarm Review dialog box is displayed.
Scrolling the screen
Not all alarm events are displayed on the current screen. To see other alarm events, press the
symbols
and
displayed on the Alarm Review dialog box.
Selecting alarm event of specific parameter
The monitor can review an alarm event of a specific parameter. To view the alarm event of the
specific parameter, select Menu > Review > Alarm Event and choose Event Type to select the
desired parameter name from the pop-up list.
Setting time index
The user can set End time of alarm review by selecting the Time Index option displayed on the
Alarm Event Review dialog box.
If the user selects Current Time on the pop-up dialog box, the alarm events occurring before the
current time are displayed on the alarm event review dialog box.
- 83 -
Instructions for Use Vista 120 Patient Monitor
If the user selects User Define, he can define the review time by setting Year, Month, Day,
Hour, Min and Sec. The alarm events occur before the User Define option is displayed on the
alarm event review dialog box.
Recording
The monitor can record the alarm event in the alarm event review window. For more information,
refer to chapter “Recording”.
Arr review
Select ECG Setup > Arr Analysis> Arr Review or Menu > Review > Arr Review to display
the Arr Review dialog box. The dialog box displays the latest arrhythmia events.
Selecting specific arrhythmia event
The monitor can review a specific arrhythmia event. To view the specific arrhythmia event, select
Menu > Review > Arrhythmia Review > Rename and choose the desired name from the
pop-up list.
Scrolling the screen
Not all arrhythmia events can be displayed on the current screen. To see the other arrhythmia
events, press the symbol
and
displayed on the Arrhythmia Review dialog box.
Deleting arrhythmia events
The user can delete the arrhythmia events displayed on the current screen by selecting Delete on
the Arrhythmia Review dialog box.
Recording
The monitor can record the waveforms of the arrhythmia events displayed on the current screen.
For more information, refer to chapter “Recording”.
- 84 -
Instructions for Use Vista 120 Patient Monitor
Calculation and titration table
The monitor provides a calculation function and titration table. Calculations are patient data that
are not directly measured but calculated by the monitor.
The monitor can perform drug calculations. Hemodynamic calculations are not described in this
Instructions for Use. To perform a calculation, select Menu > Common Function > Drug Cal.
NOTE
The drug calculation function acts only as a calculator. The patient weights in Drug
Calculation menu and in Patient Information menu are independent of each other.
Therefore changing the Weight in Drug Calculation menu will not change the weight in
the Patient Information menu.
Drug calculation
Calculation procedures
1. The drug calculation window is displayed by selecting Menu > Common Function > Drug
Calc.
2. Select the right pull-down Drug Name option and select the desired drug name among the 15
drugs which are listed as follows. The names of drug Drug A, Drug B, Drug C, Drug D and
Drug E can be defined by the user.
•
Drug A, Drug B, Drug C, Drug D and Drug E
•
AMINOPHYLLINE
•
DOBUTAMINE
•
DOPAMINE
•
EPINEPHRINE
•
HEPARIN
•
ISUPREL
•
LIDOCAINE
•
NIPRIDE
•
NITROGLYCERIN
•
PITOCIN
3. The system generates default values that cannot be treated as calculation results. The user
must input the correct parameter values.
4. Input the patient’s weight.
5. Input the correct parameter value.
6. Confirm if the calculation result is correct.
- 85 -
Instructions for Use Vista 120 Patient Monitor
Calculation unit
Each drug has the fixed unit or unit series that is used for it. Within the same unit series, the
numeric value is automatically scaled up or down as needed.
The calculation units of the drugs are as follows:
Drug
Unit
DRUG A, DRUG B, DRUG C, AMINOPHYLLINE,
g, mg, mcg
DOBUTAMINE, DOPAMINE, EPINEPHRINE,
ISUPREL, LIDOCAINE, NIPRIDE, NITROGLYCERIN
DRUG D, PITOCIN, HEPARIN
Ku, mu, Unit
DRUG E
mEq
When defining a drug, select Drug A, Drug B, Drug C, Drug D, and Drug E based on the unit
series.
NOTE
• The drug calculation is displayed as invalid prior to data entry.
•
Drip Rate and Drop Size are invalid in neonatal mode.
Titration table
After completing the drug calculation, the user can display the Titration on the Drug Calc dialog
box.
The user can change the following items in the titration table:
•
Basic
•
Step
•
Dose Type
The data in the trend table will vary with the changes above. The user can perform the following:
•
Scroll up and down the screen by selecting the symbol
graph.
•
Record the data displayed in the current dialog box by selecting the button Record.
- 86 -
and
displayed on the trend
Instructions for Use Vista 120 Patient Monitor
Strip recording
General information
A thermal dot matrix recorder is used for the monitor and can support a number of recording
types and output patient information, measurement data, and review data waveforms.
1
4
3
2
1
Start/Stop key. Press this key to start or stop recording task.
2
Recorder door
3
Paper outlet
4
Recording indicator
Performance of the recorder
•
Waveform record printed at the rate of 25 mm/s or 50 mm/s.
•
48 mm wide printout paper.
•
Record up to three waveforms.
•
User-selectable real-time recording time and waveform.
•
Auto recording interval is set by the user and the waveform displays in real time.
NOTE
You should not use the strip recorder with a low battery as recorder usage will more
quickly deplete the battery.
Recording type
The monitor provides several types of strip recordings:
•
Continuous real-time recording
- 87 -
Instructions for Use Vista 120 Patient Monitor
•
Time recording
•
Alarm recording
•
Trend graph, trend table recording
•
Arrhythmia review recording
•
Drug calculation titration recording
•
NIBP review recording
•
Alarm review recording
Starting and stopping strip recording
The strip recording can be started and stopped in the following ways:
Continuous real-time recording
Press the Start/Stop hardkey on the printer or the
Recording hardkey on the monitor to start the strip
recording, and repress it to stop the strip recording.
Auto recording
Record the three waveforms selected in the Record menu
for the duration defined in the Record menu. The strip
recording automatically stops in 8 seconds.
Alarm recording
When alarm recording is ON, it automatically starts when
alarm occurs.
Trend graph recording
Display the Trend Graph menu, and then press the Record
button in the dialog box to start strip recording.
Trend table recording
Display the Trend Table menu, then press the Record
button in the dialog box to start strip recording.
Arrhythmia review recording
Display the ECG Setup menu via shortcut key, select ARR
Analysis > ARR Review, and then press the Record button
in the dialog box to start strip recording.
Alarm review recording
Display the Alarm Review menu, then press the Record
button in the dialog box to start strip recording.
NIBP review recording
Display the NIBP Review menu, then press the Record
button in the dialog box to start recording.
Titration table recording
Display the Drug Calc menu from the Menu. Pick the
Titration button in the menu to access the Titration
window. Press the Record button to print out the currently
displayed titration table.
- 88 -
Instructions for Use Vista 120 Patient Monitor
The recorder will stop recording in the following situations:
•
The strip recording task is finished.
•
When the Start/Stop button on the front panel of the printer is pressed while the printer is
printing.
•
No paper in the recorder.
•
Because of malfunction.
Recorder operations and status messages
Record paper requirement
CAUTION
Only use standard thermo-sensitive record paper. If non-standard record paper is used,
the recorder might not function, the recording quality may be poor, and the
thermo-sensitive printhead could be damaged.
Proper operation
•
When the strip recorder is operating, do not pull the paper.
•
Do not operate the strip recorder without record paper.
Paper out
When the Recorder Out Of Paper alarm is displayed, the strip recorder will not start.
Installling paper
1. Pull out the upper arc of the strip recorder casing as shown in the following figure.
- 89 -
Instructions for Use Vista 120 Patient Monitor
2. Insert a new roll of paper into the paper cassette, printing side facing upwards.
3. Position the paper correctly.
4. Pull 2 cm of paper out and close the strip recorder casing.
NOTE
Be careful when inserting paper. Avoid damaging the thermo-sensitive print head. Do not
leave the strip recorder door open except to insert paper or troubleshoot printing
problems.
- 90 -
Instructions for Use Vista 120 Patient Monitor
Removing paper jam
When the strip recorder malfunctions, open the strip recorder door to check for a paper jam.
Remove the paper jam in the following way:
•
Cut the record paper from the feeding edge.
•
Open the strip recorder door.
•
Re-install the paper.
NOTE
• If the monitor is not installed with a strip recorder, it will indicate RECORDER SETUP
NEEDED after pressing the Record button in the dialog box.
•
Do not touch the thermo-sensitive print head when performing continuous recording.
- 91 -
Instructions for Use Vista 120 Patient Monitor
Other Functions
Nurse call
The monitor provides a dedicated nurse call port which can be connected to the hospital’s nurse
call system.
- 92 -
Instructions for Use Vista 120 Patient Monitor
Using battery
Battery power indicator
The indicator labeled Battery on the front panel of the monitor illuminates green when the
monitor is battery powered and yellow when being charged. The indicator is not illuminated
when the monitor is not powered or when using AC power.
Battery status on the main screen
The Battery Status shows the status of each battery detected and the combined battery power
remaining. It also includes an estimate of the monitoring time remaining.
Icon is shown when the battery is being charged. The yellow section represents the
battery charge level. When fully charged, the section turns green.
Icon is shown when the monitor is not equipped with a battery.
Icon is shown is there is insufficient power.
When the monitor is battery powered, the monitor switches off automatically if there is no power.
Checking battery performance
NOTE
Remove the battery from the monitor when the monitor is not used for a long periods of
time.
The performance of rechargeable batteries may deteriorate over time. The battery should be
conditioned regularly as follows:
1. Disconnect the patient from the monitor and stop all monitoring and measurement.
2. Switch the monitor power on and charge the battery for more than 6 hours continuously.
3. Disconnect the monitor and let it run until there is no battery power remaining and the
monitor shuts off.
4. Reconnect the monitor to mains.
This should improve the battery running time. If the running time is less than listed in the
specification, change the battery or contact the customer service.
WARNING
Before using the rechargeable lithium-ion battery (hereinafter called battery), be
sure to read these Instructions for Use and follow all safety precautions.
WARNING
Do not destroy, disassemble or modify the battery in any way. Do not short circuit
the battery, or its terminals, to a metal object.
- 93 -
Instructions for Use Vista 120 Patient Monitor
WARNING
Do not unplug the battery while the monitor is monitoring a patient.
WARNING
Do not heat or throw battery into a fire.
WARNING
Do not store the battery at temperatures above 60 °C.
WARNING
Do not immerse, throw, or wet battery in water/seawater.
WARNING
Use the battery only in the monitor. Do not connect battery directly to an electric
outlet or cigarette lighter charger.
WARNING
If liquid leaking from the battery gets into your eyes, do not rub your eyes. Wash
them well with clean water and go to see a doctor immediately. If liquid leaks of the
battery splash onto your skin or clothes, wash well with fresh water immediately.
WARNING
Keep away from fire immediately when leakage or foul odor is detected.
WARNING
Stop using the battery if abnormal heat, odor, discoloration, damage, deformation,
or abnormal condition is detected during use, charge, or storage. Keep it away
from the monitor.
Replacing the battery
To install or replace the battery:
Battery
door
1. Pull the battery door downwards to open it.
2. Pull the metal retainer until the battery can be removed.
- 94 -
Instructions for Use Vista 120 Patient Monitor
3. Insert the new battery into the battery compartment.
4. Pull the metal retainer downward to fix the battery and close the battery door.
Recycling the battery
When the battery no longer holds a charge, it should be replaced. Remove the old battery from
the monitor and recycle it properly.
NOTE
To prolong the life of rechargeable battery, charge it at least once every six months for a
charging time of about five hours.
- 95 -
Instructions for Use Vista 120 Patient Monitor
Care and cleaning
Use only the Dräger-approved substances and methods listed in this chapter to clean or disinfect
the equipment. The warranty does not cover damage caused by using unapproved substances or
methods.
Dräger makes no claims regarding the efficacy of the listed chemicals or methods as a means for
controlling infection. Contact infectious disease experts in your hospital for details.
General points
WARNING
Before cleaning, sterilizing, or disinfecting the monitor or any accessory, make
sure that the equipment is switched off and disconnected from the power line.
Keep the monitor, cables, and accessories free of dust and dirt.
CAUTION
To prevent the device from damage, follow this procedure:
• Always dilute cleaning, sterilizing, or disinfecting agents according to the
manufacturer’s instructions or use the lowest possible concentration.
•
Do not immerse any part of the equipment or any accessory in liquid or pour liquid
onto them.
•
Never use bleach.
•
Do not allow liquid to enter the case.
•
Never use abrasive material (such as steel wool or silver polish).
CAUTION
If you spill liquid on the equipment, battery, or accessories, or they are accidentally
immersed in liquid, contact your service personnel or DrägerService.
Cleaning
Cleaning the monitor
Regular cleaning of the monitor shell and the screen are strongly recommended. Use only
non-caustic detergents such as soap and warm water (40 °C maximum) to clean the monitor shell.
Do not use strong solvents such as acetone or trichloroethylene.
Exercise caution when cleaning the monitor screen. Do not permit liquids to enter the monitor
case. Do not allow water or cleaning solution to enter the measurement connectors.
Examples of disinfectants that can be used on the instrument casing are as follows:
•
Tenside;
•
Diluted Ammonia Water < 3%;
- 96 -
Instructions for Use Vista 120 Patient Monitor
•
Diluted Sodium Hypochlorite (Bleaching agent);
•
Diluted Formaldehyde 35 to 37 %;
•
Hydrogen Peroxide 3 %;
•
Alcohol;
•
Isopropanol.
NOTE
• The diluted sodium hypochlorite from 500 ppm (1:100 diluted bleaching agent) to
5000 ppm (1:10 bleaching agents) is very effective. The concentration of the diluted
sodium hypochlorite depends on how many organisms (blood, mucus) on the surface
of the chassis to be cleaned.
•
The monitor and sensor surface can be cleaned with hospital-grade ethanol and
allowed to air dry or dried using a clean cloth.
Cleaning the accessories
Cleaning the Cables and Lead Wires
CAUTION
Do not use acetone or ketone solvents for cleaning; do not use an autoclave or steam
cleaner.
CAUTION
The decision to sterilize must be made per your hospital’s policy with an awareness of the
effect on the integrity of the cable and leadwire.
Cables and leadwires can be cleaned with a warm, damp cloth and mild soap, or isopropyl
alcohol wipes. Refer to the previous page for safe cleaning techniques.
Cleaning the blood pressure cuff
Use a soft brush to wash the cuff surface and clean as follows:
NOTE
• Do not immerse the rubber tube or allow liquid to enter either the cuff or the tube when
washing it.
•
Do not scrub or knead the cuff.
Cleaning the CO2 sensor
Clean the sensor surface with a damp cloth. Ensure the sensor windows are clean and dry.
•
Do not immerse the sensor.
•
Do not attempt to sterilize the sensor.
- 97 -
Instructions for Use Vista 120 Patient Monitor
An adapter can be cleaned by rinsing in a warm soapy solution, followed by soaking in a liquid
disinfectant, pasteurized, or cold sterilized (glutaraldehyde). It should then be rinsed with sterile
water and dried. Before reusing any adapter, ensure the windows are clean and free of any
residue.
For other applied parts, the manufacturer must be consulted for cleaning, sterilization, or
disinfecting methods.
Other accessories
For cleaning other accessories, contact the manufacturers for details.
Sterilization
To avoid extended damage to the equipment, sterilization is only recommended when stipulated
as necessary in the hospital’s policy. Sterilization facilities should be cleaned first.
Recommended sterilization material: Ethylate and Acetaldehyde.
Appropriate sterilization materials for ECG lead and blood pressure cuff are described in those
chapters.
WARNING
Sterilize and disinfect as needed to prevent cross infection between patients.
CAUTION
Use a moistened cloth to wipe up any agent remaining on the monitor.
Disinfection
WARNING
Mixing bleach and ammonia (or other disinfecting solutions) may produce
hazardous gases.
To avoid extended damage to the equipment, disinfection is only recommended when stipulated
as necessary in the hospital’s policy. Disinfection facilities should be cleaned first.
Appropriate disinfection materials for ECG leads, SpO2 sensor, blood pressure cuff, TEMP probe,
IBP sensor are described in those chapters.
Recommended types of disinfecting agents are:
•
Alcohol
•
Aldehyde
CAUTION
Do not use etO gas or formaldehyde to disinfect the monitor.
- 98 -
Instructions for Use Vista 120 Patient Monitor
Maintenance
WARNING
Failure on the part of the responsible institution to implement a satisfactory
maintenance schedule may cause undue equipment failure and possible health
hazards.
WARNING
If the monitor is mechanically damaged, or if it is not working properly, do not use
it. Contact your technical personnel.
WARNING
If you discover a problem with any of the equipment, contact your service
personnel or your authorized supplier.
Inspecting
The overall check of the monitor, including the safety check, should be performed by qualified
personnel every 24 months, and following repair.
The following items should be checked:
•
If the environmental condition and power supply meet requirements.
•
If the power supply cord is damaged and the insulation is not cut.
•
If the device and accessories are damaged.
•
Specified accessories.
•
If the alarm system works properly.
•
If the recorder works properly and there is sufficient paper.
•
Battery performance
•
If all monitoring functions are in good condition.
•
If grounding resistance and leakage function properly.
If any damage or abnormality is found, do not use the monitor and contact customer support.
Maintenance task and test schedule
The following maintenance tasks are for Dräger-qualified service professionals only. Contact a
Dräger-qualified service provider if the monitor needs a safety or performance test. Clean and
disinfect equipment before testing or maintaining it.
Maintenance and Test Schedule
Frequency
Safety checks. Selected tests on
the basis of IEC 62353
At least once every two years, or as needed, following
any repairs where the power supply is removed or
replaced, or if the monitor has been dropped.
- 99 -
Instructions for Use Vista 120 Patient Monitor
Maintenance and Test Schedule
Frequency
Check ECG synchronization of the At least once every two years, or as needed.
monitor
NIBP leakage inspection and
calibrations
At least once every two years, or as specified by local
laws.
CO2 calibration and performance
test
At least once every two years, or if you suspect the
measurement values are incorrect.
Replacing fuse
To replace the blown fuse:
1. Switch off the monitor and unplug the power cable.
2. Carefully place the monitor upside down on a flat surface covered with cloth or other
protecting pad.
3. Pull ot the fuse box from the groove and you will see the fuse.
4. Remove the old fuse. Take out a new fuse with the following specifications:
Ф 5 mm x 20 mm: Model: T1.6AL 250 V.
5. Place the new fuse in the fuse socket and push the box back into position.
- 100 -
Instructions for Use Vista 120 Patient Monitor
Disposal
CAUTION
The device and its components must be disinfected and cleaned before disposal!
For countries subject to EU Directive 2002/96/EC:
This device is subject to EU directive 2002/96/EC (WEEE). In order to comply with its
registration according to this directive, it may not be disposed of at municipal collection
points for waste electrical and electronic equipment. Dräger has authorized a company to
collect and dispose of this device. To initiate take-back or for further information, visit us
on the Internet at www.draeger.com and navigate to the DrägerService area where you
will find a link to “WEEE”. If you have no access to our website, contact your loacal
Dräger organization.
- 101 -
Instructions for Use Vista 120 Patient Monitor
Service policy
Repair of the device may only be carried out by trained service personnel otherwise the correct
functioning of the device may be compromised.
Regular bi-annual maintenance (functional and safety test) according to IEC 62353 is
recommended, in addition to national regulations and laws (for example, accident prevention
regulations).
Connecting the medical device to other medical devices could result in additional maintenance
requirements.
Consult the documentation for these other devices to identify additional requirements.
Dräger recommends contracting with DrägerService for any repairs. Use only authentic Dräger
repair parts during maintenance. Using non-Dräger repair parts may adversely affect the
operation of the device.
If the product fails to function properly, or if assistance is needed, for service and for spare parts,
contact Dräger’s service center. A representative will assist in troubleshooting the problem and
will make every effort to solve it over the phone or e-mail, avoiding potential unnecessary
returns.
In case a return can not be avoided, the representative will record all necessary information and
will provide a Return Material Authorization (RMA) form that includes the appropriate return
address and instructions. An RMA form must be obtained prior to any return.
If you have any questions about maintenance, technical specifications, or malfunctions of devices,
contact your local distributor.
- 102 -
Instructions for Use Vista 120 Patient Monitor
Accessories
Accessories can be ordered from your local Dräger representative.
WARNING
Never reuse disposable transducers, sensors or other accessories that are
intended for single-use or single patient use only. Reuse may compromise device
functionality and system performance and cause a potential hazard.
WARNING
Only use accessories that are approved by Dräger. Using accessories that are not
approved by Dräger may compromise device functionality and system
performance and cause a potential hazard.
WARNING
Do not use a sterilized accessory if its packaging is damaged.
NOTE
Transducers and sensors have a limited shelf life. Refer to the package labeling.
The following cables may not all be available in all countries. Check availability with your local
Dräger supplier.
ECG accessories
The following table lists the optional configuration for the monitor:
Part Number
Accessory
2606185
ECG trunk cable, 5-lead, Defib, IEC, 2.6 m, reusable
2606186
ECG limb wires, 5-lead, snap, IEC, 0.9 m, reusable
2606187
ECG limb wires, 5-lead, clip, IEC, 0.9 m, reusable
2606192
ECG cable, 5-lead, snap, Defi, AHA, 3.5 m, reusable
2606193
ECG cable, 5-lead, clip, AHA, 3.5 m, reusable
2606195
ECG cable, 3-lead, snap, AHA, 3.5 m, reusable
2606196
ECG trunk cable, 3-lead, Defib, IEC, 2.6 m, reusable
2606197
ECG limb wires, 3-lead, clip, IEC, 0.9 m, reusable
2606245
ECG Electrodes, adult, disposable, 30 pieces
2606200
ECG Electrodes, adult, disposable, 100 pieces
2606246
ECG Electrodes, child, neo disposable, 50 pieces
- 103 -
Instructions for Use Vista 120 Patient Monitor
SpO2 accessories
Part Number
Accessory
2606201
SpO2 Finger Sensor, adult, 2.5 m, reusable
2606202
SpO2 Finger Sensor, adult, 1 m, reusable
2606205
SpO2 Silicone Soft-tip Sensor, adult, 1 m, reusable
2606206
SpO2 Silicone Soft-tip Sensor, pediatric, 1 m, reusable
2606207
SpO2 Extension cable, 2m
2606208
SpO2 Sensor, adult, 0.5 m, disposable, 25 pieces
2606209
SpO2 Sensor, pediatric, 0.5 m, disposable, 25 pieces
2606210
SpO2 Sensor, Infant, 0.5 m, disposable, 25 pieces
2606211
SpO2 Sensor, Neonate, 0.5 m, disposable, 25 pieces
NIBP accessories
Part Number
Accessory
2606212
NIBP Cuff, Adult, 25 cm – 35 cm, reusable
2606213
NIBP Cuff, Larger Adult, 33 cm – 47 cm, reusable
2606214
NIBP Cuff, Child, 18 cm – 26 cm, reusable
2606215
NIBP Cuff, Infant, 10 cm – 19 cm, reusable
2606271
NIBP Cuff, Neonatal #1, 3 cm – 6 cm, disposable
2606272
NIBP Cuff, Neonatal #2, 4 cm – 8 cm, disposable
2606273
NIBP Cuff, Neonatal #3, 6 cm – 11 cm, disposable
2606274
NIBP Cuff, Neonatal #4, 7 cm – 13 cm, disposable
2606275
NIBP Cuff, Neonatal #5, 8 cm – 15 cm, disposable
2606218
NIBP Tube, 3 m
2606219
NIBP Tube, for neonatal cuff, 3 m
- 104 -
Instructions for Use Vista 120 Patient Monitor
Temp accessories
Part Number
Accessory
2606220
Temperature Probe Skin, adult, 3 m, reusable
2606221
Temperature Probe, rectal / oral, 3 m, reusable
IBP accessories
Part Number
Accessory
2606224
IBP Pressure transducer interface cable, BD
2606225
IBP Pressure transducer kit, BD, disposable, 5 pieces
CO2 accessories
Part Number
Accessory
2606226
EtCO2 Module with disp. Adult airway adapter
2606227
CO2 Airway Adapter, Adult, disposable, 10 pieces
2606228
CO2 Airway Adapter, Neonatal (infant/pediatric), 10 pcs.
Device accessories
Part Number
Accessory
2606229
Vista 120 Rechargeable Lithium-Ion Battery
2606231
Vista 120 Printing paper, 4 rolls
2606232
Vista 120 Ground Cable
2606233
Vista 120 Mounting adaptor
Spare parts
Part Number
Accessory
2606229
Vista 120 Rechargeable Lithium-Ion Battery
2606260
Fuse
- 105 -
Instructions for Use Vista 120 Patient Monitor
Technical Data
Classification
Protection class
Class I equipment and internal powered equipment
EMC class
Group I, Class A
Degree of protection against
electric shock
CF: ECG (RESP), TEMP, IBP
BF: SpO2, NIBP, CO2
Liquid ingress protection
IPX1
Disinfection/sterilization method Refer to chapter “Care and Cleaning” for details.
Mode of operation
Continuous
Device specifications
Size
408 mm (L) × 157 mm (W)× 316 mm (H)
Weight
7 kg
Ambient conditions
Temperature
Operating
+5 to +40 °C
Transport and storage
-20 to +55 °C
Relative humidity
Operating
25 to 80 % (non-condensing)
Transport and storage
25 to 93 % (non-condensing)
Altitude
Operating
860 to 1060 hPa
Transport and storage
700 to 1060 hPa
Leakage current
Applied part
Normal condition
Single fault condition
Earth leakage current
<0.5 mA
<1 mA
Enclosure leakage current
<0.1 mA
<0.5 mA
CF
AC: <0.01 mA
DC: <0.01 mA
AC: <0.05 mA
DC: <0.05 mA
BF
AC: <0.1 mA
AC: <0.5 mA
Patient leakage current
- 106 -
Instructions for Use Vista 120 Patient Monitor
DC: <0.01 mA
Patient leakage current
(Mains on applied parts)
Patient auxiliary current
DC: <0.05 mA
CF
< 0.05 mA
BF
< 5 mA
CF
AC: <0.01 mA
DC: <0.01 mA
AC: <0.05 mA
DC: <0.05 mA
BF
AC: <0.1 mA
DC: <0.01 mA
AC: <0.5 mA
DC: <0.05 mA
Power supply
Power supply
100 to 240 V~, 50 Hz/60 Hz
Pmax = 110 VA
FUSE T 1.6 AL, 250 V
Battery (optional)
Quantity
1
Capacity
4.2 Ah
Battery life
120 min (At 25 °C, continuous SpO2 measuring mode and
NIBP automatic measuring mode)
Battery charge time
320 min (Monitor is on or in standby mode.)
Display
Display
Messages
Display screen: 15” color TFT
A maximum of 11 waveforms
Resolution: 1024 × 768
One power LED
Two alarm LEDs
One charge LED
Recorder
Record width
48 mm
Paper speed
25 mm/s, 50 mm/s
Trace
Up to 3
- 107 -
Instructions for Use Vista 120 Patient Monitor
Recording types
8 second real-time recording
Auto 8 second recording
Parameter alarm recording
Trend recording
Titration table recording
Review
Trend review
Short
1 hr, 1 sec. resolution
Long
120 hrs, 1 min. resolution
Review
1200 sets NIBP measurement data
ECG
3-, 5-lead monitoring
Lead mode
3-Lead: I, II, III
5-Leads: I, II, III, aVR, aVL, aVF, V
Waveform
3-Lead: 1-channel waveform
5-Lead: 2-channel waveform, max. seven waveforms;
Lead naming style
AHA, IEC
Display sensitivity
1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV
(×0.5), 10 mm/mV (×1), 20 mm/mV (×2), AUTO gain
Sweep
12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth (-3dB)
Diagnosis: 0.05 to 150 Hz
Monitor: 0.5 to 40Hz
Surgery: 1 to 20Hz
CMRR (Common Mode
Rejection Ratio)
Diagnosis: >95 dB (the notch filter is off)
Monitor: >105 dB (the notch filter is on)
Surgery: >105 dB (the notch filter is on)
Notch
50 Hz/60 Hz (notch filter can be selected manually)
Differential input impedance
>5 MΩ
Input signal range
±8 mVPP
- 108 -
Instructions for Use Vista 120 Patient Monitor
Electrode offset potential
Tolerance
±500 mV
Auxiliary current (Leads off
detection)
Active electrode: <100 nA
Input offset current
≤0.1 μA
Recovery time after
defibrillation
<5 s
Patient leakage current
<10 μA (normal condition)
Scale signal
1 mVPP, accuracy is ±5 %
System noise
<30 μVPP
ESU protection
Cut mode: 300 W
Reference electrode: <900 nA
Coagulation mode: 100 W
Recovery time: ≤10 s
Meets the requirements of ANSI/AAMI EC13-2002:
Sect. 4.1.2.1 a)
ESU noise suppression
Tested according to the test method in EC13: 2002
Sect.5.2.9.14, it complies with the standard.
Pace pulse
Pulse indicator
Pulses are marked if the requirements of ANSI/AAMI
EC13:2002, Sect. 4.1.4.1 are met:
Amplitude: ±2 to ±700 mV
Width: 0.1 to 2 ms
Rise time: 10 to 100 μs
Pulse rejection
For heart rate meter, pulse is rejected if the requirements of
ANSI/AAMI EC13:2002, Sect. 4.1.4.1 are met:
Amplitude: ±2 to ±700 mV
Width: 0.1 to 2 ms
Rise time: 10 to 100 μs
Heart rate
Range
ADU: 15 to 300 bpm
PED/NEO: 15 to 350 bpm
Accuracy
±1 % or ±1 bpm, whichever is greater
Resolution
1 bpm
- 109 -
Instructions for Use Vista 120 Patient Monitor
Sensitivity
≥300 μVPP
PVC
Range
ADU: 0 to 300 PVCs/ min
PED/NEO: 0 to 350 PVCs/ min
Resolution
1 PVCs/min or 2% of measurement, whichever is greater
ST value
Range
-2.0 to +2.0 mV
Accuracy
The max. of ±0.02 mV or 10% (-0.8 to +0.8 mV)
Resolution
0.01 mV
HR averaging method
Method 1
Normally, heart rate is computed by averaging the 12 most
recent RR intervals.
Method 2
If each of three consecutive RR intervals is greater than
1200 ms, then the four most recent RR intervals are
averaged to compute the HR.
Range of sinus and SV rhythm
Brady
ADU: 120 to 300 bpm
PED/NEO: 160 to 350 bpm
Normal
ADU: 41 to 119 bpm
PED/NEO: 61 to 159 bpm
Tachy
ADU: 15 to 40 bpm
PED/NEO: 15 to 60 bpm
Range of ventricular rhythm
Ventricular tachycardia
The interval of 5 consecutive ventricular beats is less than
600 ms
Ventricular rhythm
The interval of consecutive ventricular beats ranges from
600 ms to 1000 ms
Ventricular bradycardia
The interval of 5 consecutive ventricular beats is more than
1000 ms
Time to alarm for tachycardia
Ventricular tachycardia
Gain 1.0: 10 s
1 mV 206bpm
Gain 0.5: 10 s
Gain 2.0: 10 s
- 110 -
Instructions for Use Vista 120 Patient Monitor
Ventricular tachycardia
Gain 1.0: 10 s
2 mV 195 bpm
Gain 0.5: 10 s
Gain 2.0: 10 s
Response time of heart rate
meter to change in HR
HR range: 80 to 120 bpm
Range : 7 to 8 s, average is 7.5 s
HR range: 80 to 40 bpm
Range : 7 to 8 s, average is 7.5 s
Tall T-wave rejection
Exceeds ANSI/AAMI EC13:2002, Sect. 4.1.2.1 c)
minimum recommended 1.2 mV T-Wave amplitude
Accuracy of heart rate meter and
response to irregular rhythm
According with ANSI/AAMI EC13:2002, Sect.4.1.2.1 e)
The HR value after 20s:
Ventricular bigeminy: 80 ±1 bpm
Slow alternating ventricular bigeminy: 60 ±1 bpm
Rapid alternating ventricular bigeminy: 120 ±1 bpm
Bidirectional systoles: 90 ±1 bpm
16 different arrhythmia analyses
Non-paced patient
Paced patient
ASYSTOLE
R on T
ASYSTOLE
VFIB/VTAC
PVC
TACHY
COUPLET
TACHY
BRADY
VT>2
BRADY
PNC
BIGEMINY
MISSED BEATS
PNP
TRIGEMINY
VWNT
IRR
VBRADY
RESP
Method
Impedance between RA-LL, RA-LA
Baseline impedance range
200 to 2500 Ω (cable resistance = 0 K)
2200 to 4500 Ω (leads cables 1KΩ resistance)
Measuring sensitivity
0.3 Ω (baseline impedance 200 to 4500 Ω)
Noise
< 0.1 Ω (3, 5-lead monitoring)
Max. dynamic range
500 Ω resistance, 3 Ω variable resistance, no clipping
Waveform bandwidth
0.2 to 2.5 Hz (-3 dB)
- 111 -
Instructions for Use Vista 120 Patient Monitor
RR measuring range:
Adult
0 to 120 rpm
Neo/Ped
0 to 150 rpm
Resolution
1 rpm
Accuracy
±2 rpm
Gain selection
×0.25, ×0.5, ×1, ×2, ×3, ×4, ×5
Respiration excitation
waveform
< 400 μA, sinusoid, 62.8 kHz (± 10%)
NIBP
Method
Oscillometric
Mode
Manual, Auto, Continuous
Measuring interval in AUTO
mode
1/2/3/4/5/10/15/30/60/90/120/240/480 min
Continuous
5 min, interval is 5 s
Measuring type
Systolic Pressure, Diastolic Pressure, Mean Pressure
Alarm type
SYS, DIA, MAP
Measuring and alarm rang
Adult mode
SYS: 40 to 270 mmHg
DIA: 10 to 215 mmHg
MAP: 20 to 235 mmHg
Pediatric mode
SYS: 40 to 200 mmHg
DIA: 10 to 150 mmHg
MAP: 20 to 165 mmHg
Neonatal mode
SYS: 40 to 135 mmHg
DIA: 10 to 100 mmHg
MAP: 20 to 110 mmHg
Cuff pressure measuring
range
0 to 300 mmHg
Pressure resolution
1 mmHg
Maximum standard
deviation
8 mmHg
- 112 -
Instructions for Use Vista 120 Patient Monitor
Maximum measuring period
Adult/Pediatric
120 s
Neonate
90 s
Typical measuring period
30 to 45 s (depend on HR/motion disturbance)
Overpressure protection (Dual overpressure protection)
Adult
297 ±3 mmHg
Pediatric
240 ±3 mmHg
Neonatal
147 ±3 mmHg
Measuring range
40 to 240 bpm
Accuracy
±3 or 3.5%, whichever is larger
PR
SpO2
Measuring Range
0 to 100 %
Alarm Range
0 to 100 %
Resolution
1%
Accuracy
Adult (including Pediatric)
±2 % (70 to 100 % SpO2)
Undefined (0 to 69 % SpO2)
Neonate
±3 % (70 to 100 % SpO2)
Undefined (0 to 69 % SpO2)
Pulse Rate
Measuring range
25 to 300 bpm
Alarm range
30 to 300 bpm
Accuracy
±2 bpm
Data update period
1s
Wave length
Red light
660±3 nm
Infrared light
905±5 nm
- 113 -
Instructions for Use Vista 120 Patient Monitor
TEMP
Channel
2
Measuring range
0 to 50 °C
Alarm range
0 to 50 °C
Sensor type
YSI-10K
Resolution
+0.1 °C
Accuracy (without sensor)
±0.1 °C
Refresh time
Every 1 to 2 s
IBP
Pressure measuring range
-50 to +300 mmHg
Resolution
1 mmHg
Accuracy (without sensor)
± 2 % or ±1 mmHg, whichever is larger
Pressure sensor
Sensitivity
5 (μV/V/mmHg)
Impedance
300 to 3000 Ω
Frequency response
d.c. to 12.5 or 40 Hz
Zero
Range: ±200 mmHg
Accuracy: ±1 mmHg
Measuring and alarm range
Art
0 mmHg to 300 mmHg
PA
-6 to +120mmHg
CVP/RAP/LAP/ICP
-10 to +40 mmHg
P1/P2
-50 to +300 mmHg
Volume displacement of MSI
4.5 x 10-4 in3 / 100 mmHg
CO2
Method
Infra-red Absorption Technique
Unit
mmHg, %, kPa
Measuring range
EtCO2
0 to 150 mmHg
- 114 -
Instructions for Use Vista 120 Patient Monitor
FiCO2
3 to 50 mmHg
AwRR
0 to 150 rpm (MainStream)
Resolution
EtCO2
1 mmHg
FiCO2
1 mmHg
AwRR
1 rpm
EtCO2 Accuracy
± 2 mmHg, 0 to 40 mmHg
± 5 % of reading, 41 to 70 mmHg
± 8 % of reading, 71 to 100 mmHg
± 10 % of reading, 101 to 150 mmHg
AwRR Accuracy
± 1 rpm
Apnea Alarm Delay
10 s, 15 s, 20 s (default), 25 s, 30 s, 35 s, 40 s
Calculation Method
BTPS (Body Temperature Pressure Saturated)
Sample Gas Flowrate
50 ml/min
Stability
Short Term Drift
Drift over 4 hours < 0.8 mmHg
Long Term Drift
120 hour period
O2 Compensation
Range
0 to 100 %
Resolution
1%
Default
16 %
Response time
60 ms
- 115 -
Instructions for Use Vista 120 Patient Monitor
EMC Declaration
General information
The EMC compliance of the medical device is also applicable to the external cables, transducers,
and accessories specified in the list of accessories. In addition, accessories which do not affect
EMC compliance may be used if no other reasons forbid their use (see other sections of the
Instructions for Use). The use of non-compliant accessories may result in increased emissions or
decreased immunity of the medical device.
The medical device may only be used adjacent to or stacked with other devices when the
configuration is approved by Dräger. If adjacent or stacked use of non-approved configurations is
inevitable, the medical device must be observed to verify normal operation in the said
configuration. In any case, strictly observe the Instructions for Use of the other devices.
Electromagnetic emissions
When selecting wireless systems (wireless communication devices, pager systems, etc.) for use in
installations where wireless networking is used, ensure that operating frequencies are compatible.
Low-level signals such as ECG are particularly susceptible to interference from electromagnetic
energy. While the equipment meets the testing described below, it will not ensure perfect
operation. The ‘quieter’ the electrical environment the better. In general, increasing the distance
between electrical devices decreases the likelihood of interference.
Electromagnetic environment
The medical device is intended for use in an electromagnetic environment as specified below.
The user must ensure its use in such an environment.
Emissions
Radio frequency emissions
(CISPR 11)
Compliance according to
Group 1
Class A
Class A
Harmonic emissions
(IEC 61000-3-2)
Voltage fluctuations/flicker Complies
emissions (IEC 61000-3-3)
- 116 -
Electromagnetic environment
The medical device uses RF energy
only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
The medical device is suitable for
use in all establishments other than
domestic establishments and those
directly connected to the public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
Instructions for Use Vista 120 Patient Monitor
Electromagnetic immunity
The medical device is intended for use in an electromagnetic environment as specified below.
The user must ensure its use in such an environment.
Immunity against
Electrostatic
discharge (ESD)
(IEC 61000-4-2)
Electrical fast
transients/ bursts
(IEC 61000-4-4)
Surge on AC mains
lines (IEC
61000-4-5)
Power frequency
magnetic field (50/60
Hz) (IEC 61000-4-8)
Test level
(IEC 60601-1-2)
Contact discharge:
±6 kV
Air discharge:
±8 kV
Compliance level
Power supply lines:
±2 kV
Longer input / output
lines: ±1 kV
Common mode:
±2 kV
Differential mode:
±1 kV
3 A/m
Voltage dips and
short interruptions on
AC mains input lines
(IEC 61000-4-11)
Dip >95 %, 0.5
periods
Dip 60 %, 5 periods
Dip 30 %, 25 periods
Dip >95 %, 5 seconds
Radiated RF (IEC
61000-4-3)
80 MHz to 2.5 GHz:
3 V/m
Electromagnetic
environment
Contact discharge:
Floors should be wood,
±6 kV
concrete or ceramic tile.
If floors are covered
Air discharge:
with synthetic material,
±8 kV
the relative humidity
should be at least 30 %.
Power supply lines: Mains power quality
±2 kV
should be that of a
Longer input / output typical commercial or
hospital environment.
lines: ±1 kV
Common mode:
Mains power quality
±2 kV
should be that of a
typical commercial or
Differential mode:
hospital environment.
±1 kV
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Dip >95 %, 0.5 Mains power quality
periods
should be that of a
typical commercial or
Dip 60 %, 5 periods
hospital environment. If
Dip 30 %, 25 periods
Dip >95 %, 5 seconds the user of the medical
device requires
continued operation
during mains power
supply interruptions, it
is recommended that
the medical device be
powered from an
uninterruptible power
supply or a battery.
3 V/m
Portable and mobile RF
communications
equipment should be
used no closer to any
part of Vista 120,
including cables, than
the recommended
- 117 -
Instructions for Use Vista 120 Patient Monitor
Immunity against
Conducted RF (IEC
61000-4-6)
Test level
(IEC 60601-1-2)
150 kHz to 80 MHz:
3 VRMS
Compliance level
3 VRMS
Electromagnetic
environment
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
⎡ 3 .5 ⎤
d =⎢ ⎥ P
⎣ V1 ⎦
⎡ 3 .5 ⎤
d = ⎢ ⎥ P 80 MHz
⎣ E1 ⎦
to 800 MHz
⎡7⎤
d = ⎢ ⎥ P 800
⎣ E1 ⎦
MHz to 2.5 GHz
Where P is the
maximum output power
rating of the transmitter
in watts (W) according
to the transmitter
manufacturer and d is
the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey1, should be less
than the compliance
level in each frequency
range2.
Interference may occur
in the vicinity of
equipment marked with
the following symbol:
- 118 -
Instructions for Use Vista 120 Patient Monitor
Immunity against
Test level
(IEC 60601-1-2)
Compliance level
Electromagnetic
environment
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which Vista 120 is used
exceeds the applicable RF compliance level above, Vista 120 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating Vista 120.
2
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
- 119 -
Instructions for Use Vista 120 Patient Monitor
Recommended separation distances to portable and mobile RF
telecommunication devices
The separation distances below are in accordance with IEC 60601-1-2.
Rated maximum
output power of
transmitter
Separation distance according to frequency of transmitter(m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
(W)
⎡ 3 .5 ⎤
d =⎢ ⎥ P
⎣ V1 ⎦
⎡ 3 .5 ⎤
d =⎢ ⎥ P
⎣ E1 ⎦
⎡7⎤
d =⎢ ⎥ P
⎣ E1 ⎦
0.01
0.12
0.12
0.23
0.1
0.36
0.37
0.74
1
1.16
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
- 120 -
Instructions for Use Vista 120 Patient Monitor
Default settings
This appendix documents the most important default settings of the monitor as it is delivered
from the factory.
Note: If the monitor has been ordered preconfigured to your requirements, the settings at delivery
will be different from those listed here.
Patient information default settings
Patient information settings
Patient type
Adult
Pace
Off
Alarm default settings
Alarm settings
Pause time
120 s
Alarm Mute
On
Sensor off alarm
On
Alarm latch
Off
ECG default settings
ECG settings
ADU
PED
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Alarm high limit
120
160
200
Alarm low limit
50
75
100
Pace
Off
Lead type
5 Leads
Display
Normal
Filter
Monitor
Smart lead off
Off
Heart volume
3
- 121 -
NEO
Instructions for Use Vista 120 Patient Monitor
ST analysis
ADU
ST analysis
On
Alarm switch
Off
Alarm level
Medium
Alarm record
Off
Alarm high limit
(ST-X)
0.2
Alarm low limit
(ST-X)
-0.2
PED
NEO
X stands for I, II, III, aVR, aVL, aVF, V.
ARR analysis
ARR analysis
Off
PVCs alarm level
Medium
PVCs alarm switch
Off
PVCs alarm record
Off
ARR alarm settings
Alarm switch
Alarm level
Alarm record
ASYSTOLE
On
Medium
Off
VFIB/VTAC
On
Medium
Off
R ON T
On
Medium
Off
VT > 2
On
Medium
Off
COUPLET
On
Medium
Off
PVC
On
Medium
Off
BIGEMINY
On
Medium
Off
TRIGEMINY
On
Medium
Off
TACHY
On
Medium
Off
BRADY
On
Medium
Off
MISSEDBEATS
On
Medium
Off
IRR
On
Medium
Off
PNC
On
Medium
Off
PNP
On
Medium
Off
VBRADY
On
Medium
Off
VENT
On
Medium
Off
- 122 -
Instructions for Use Vista 120 Patient Monitor
RESP
RESP settings
ADU
PED
NEO
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Alarm high limit
30
30
100
Alarm low limit
8
8
30
Apnea time
20 s
Calculation type
Auto
Resp hold type
Auto
Sweep
12.5 mm/s
Amplitude
1
PED
NEO
SpO2
SpO2 settings
ADU
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Alarm high limit
100
100
95
Alarm low limit
90
90
88
Pitch tone
Off
Sweep
12.5 mm/s
PED
NEO
PR
PR settings
ADU
PR source
SpO2
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Alarm high limit
120
160
200
Alarm low limit
50
75
100
Pulse volume
3
- 123 -
Instructions for Use Vista 120 Patient Monitor
Alarm source
Auto
NIBP
NIBP settings
ADU
PED
NEO
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Alarm high limit (SYS)
160
120
90
Alarm low limit (SYS)
90
75
40
Alarm high limit (Mean)
110
90
70
Alarm low limit (Mean)
60
50
30
Alarm high limit (Dia)
90
70
60
Alarm low limit (Dia)
50
40
20
Inflation value
160
140
100
Unit
mmHg
Interval
Manual
PED
NEO
TEMP
TEMP settings
ADU
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Alarm high limit (T1)
39.0
39.0
39.0
Alarm low limit (T1)
36.0
36.0
36.0
Alarm high limit (T2)
39.0
39.0
39.0
Alarm low limit (T2)
36.0
36.0
36.0
Alarm high limit (TD)
2.0
2.0
2.0
Unit
°C
- 124 -
Instructions for Use Vista 120 Patient Monitor
IBP
IBP settings
ADU
PED
NEO
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Unit
mmHg
Filter
12.5 Hz
SYS, DIA,
MEAN
SYS, DIA, MEAN
SYS, DIA, MEAN
Alarm high limit
(ART, P1, P2)
160, 90, 110
160, 90, 110
160, 90, 110
Alarm low limit
(ART, P1, P2)
90, 50, 70
90, 50, 70
90, 50, 70
Alarm high limit (PA)
35, 16, 20
35, 16, 20
35, 16, 20
Alarm low limit (PA)
10, 0, 0
10, 0, 0
10, 0, 0
MEAN
MEAN
MEAN
Alarm high limit
(CVP, RAP, LAP, ICP)
10
10
10
Alarm low limit
(CVP, RAP, LAP, ICP)
0
0
0
CO2
CO2 settings
ADU
PED
Alarm switch
On
Alarm record
Off
Alarm level
Medium
Work mode
Measure
Unit
mmHg
Apnea time
20 s
O2 compensate
16 %
Anes agent
0%
Alarm high limit (EtCO2)
50
50
45
Alarm low limit (EtCO2)
15
20
30
- 125 -
NEO
Instructions for Use Vista 120 Patient Monitor
Alarm high limit (FiCO2)
4
4
4
Alarm high limit (AWRR)
30
30
100
Alarm low limit (AWRR)
8
8
30
Sweep
12.5 mm/s
Amplitude
Low
- 126 -
Instructions for Use Vista 120 Patient Monitor
- 127 -
Instructions for Use Vista 120 Patient Monitor
Passwords
The following dialog boxes are protected by a password to prevent unauthorized changes:
•
•
User Maintain
Demo Mode
The password can be found on this page of these Instructions for Use. Cut out this section and
store it in a safe place to keep it from unauthorized persons.
If the section with the password has been removed, ask the person responsible for the device
about the possibility of making changes in the above mentioned dialog boxes.
If the password is lost, contact DrägerService.
Passwords for Vista 120
Cut out this section and store it in a safe place to keep it from unauthorized persons.
The following dialog boxes are protected by a password to prevent unauthorized changes:
Dialog box
Password
Menu > Maintenance > User Maintain
ABC
Menu > Common Function > Demo Mode
3045
- 128 -
These instructions for use apply only to
Vista 120 SW 1.n
with the Serial No.:
If no Serial No. has been filled in by Dräger,
these Instructions for Use are provided for
general information only and are not intended
for use with any specific machine or unit.
This document is provided for customer
information only, and will not be updated or
exchanged without customer request.
Directive 93/42/EEC
concerning Medical Devices
Manufacturer:
Dräger Medical GmbH
Moislinger Allee 53-55
D-23542 Lübeck
Germany
+49 451 8 82-0
FAX +49 451 8 82-20 80
http://www.draeger.com
2606251 – GA 6498.020 enUS
© Dräger Medical GmbH
Edition: 3 – 2011-06
(Edition: 1 – 2010-12)
Dräger reserves the right to make modifications
to the equipment without prior notice.
Download