IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 39, NO. IO, OCTOBER 1992 99I Estimation of the Spatio-Temporal Correlations of Biological Electrical Sources from Their Magnetic Fields Warren E. Smith, Member, IEEE Abstract-Quasi-static electromagnetic systems, such as those found in biological systems, produce electric and magnetic fields whose temporal and spatial correlations reflect the source correlations in a straightforward manner. These fields can be noninvasively measured, providing information about the coherence properties of the source, which may directly represent ordered physiological processes of the organism. The description “biocoherence” will be adopted here to refer to the manifestation of the coherence in the magnetic measurements of these sources due solely to physiological processes. In this paper a general formulation linking the spatial and temporal coherence of measurable magnetic fields with the corresponding spatial and temporal coherence of the inaccessible current sources is derived in the quasi-static model. A method for reconstructing the spatial and temporal coherence of the source distribution is then presented. Such coherence maps would be useful descriptors of physiological processes occurring over time and space, and would represent more information than an image of the current sources frozen in time, or even a temporal sequence of such images. INTRODUCTION EN one structure of the nervous system is active or suppressed at time t , , another structure’s activity at another spatial location may be directly affected at some later time t 2 . The cause-effect relationship may be mediated by a direct connection or by some intermediate structures. A relatively low-level example of such a process is the cortical and motor rhythms taking place during walking. These are periodic activities of all of the structures involved (e.g., cortex, thalamus, cerebellum, brainstem, muscles of the body), each structure performing its own periodic function with varying amplitudes and phases, contributing to the overall coordinated effect. Another possible example is the evoked response in the cortex to a sight, sound, or touch sensation. Many areas of the cortex and subcortex may take part at different times in response to the stimulus. By examining the types of spatio-temporal correlations that normally exist between structures during a specific task, two things can be learned. First, any hidden corre- w Manuscript received July 13, 1990; revised August 27, 1991. This work was supported by a Public Health Service Biomedical Research Support Grant and the Joint Services Optics Program Grant #DAAL 03-88-K-0182. The author is with the Institute of Optics, University of Rochester, Rochester, NY 14627. IEEE Log Number 9202572. lation (in space and time) between structures, whether by direct or indirect linkage, might become evident. Second, pathology of the structures may be detectable as perturbations from the norm of the correlations. Spatio-temporal correlation maps will contain much more information than an image of source activity at a fixed time, or even a movie of the source activity over several time frames. In such movies the human observer qualitatively forms temporal correlations of the spatial processes, but may miss subtle correlations widely separated in time. For spatio-temporal processes such as nervous-system function the determination of the source correlations, and not just source snapshots, is an important added dimension of analysis. How can such spatio-temporal processes be measured? Systems that image anatomy in vivo, such as X-ray computed tomography (CT) and magnetic resonance imaging (MRI), have spatial resolution in the millimeter range and temporal resolution that is approaching tens of milliseconds, but they do not measure function. In nuclear medicine [positron emission tomography (PET) and single photon emission computed tomography (SPECT)] it is possible to measure physiological function with spatial resolutions on the order of 5-10 mm with a temporal resolution on the order of many seconds or tens of minutes, respectively , but this temporal resolution is much larger than the time constants of the active processes. Electrical potential measurements (electroencephalography, electrocardiography) [ 11 detect, with a temporal resolution on the order of milliseconds, physiological processes that produce global potential differences on the surface of the body. The spatial resolution of the sources has been poor (many centimeters), however, due to the smearing of the electric fields by the conductivities of the tissues of the body. More recently magnetic measurements have been made of the electrical activity of the body [2]-[4] using superconducting quantum interference devices (SQUID’S) [ 5 ] . The production of magnetic fields by the electrical processes within living tissue is called biomagnetism. The magnetic fields can be measured with the excellent temporal resolution of the electrical-potential measurements, but there is evidence that magnetic measurements may offer better spatial localization of the electrical sources than potential measurements [6]. Thus, biomagnetism may be 0018-9294/92$03.00 0 1992 IEEE 998 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 39, NO. 10. OCTOBER 1992 able to provide us with a unique capability impossible with other imaging modalities: a totally noninvasive data-acquisition technique that measures the spatio-temporal information of nervous-system function. Magnetic fields produced by the currents found in the brain and heart have been measured with the techniques of magnetoencephalography and magnetocardiography, respectively. One approach to source reconstruction has been to fit only a single current dipole to the measured magnetic fields. This is particularly useful in isolating epileptogenic foci of the brain. More recently interest has been developing in forming three-dimensional images of large numbers of source current dipoles [7]-[14]. The SQUID hardware has also been evolving to facilitate more simultaneous magnetic measurements. Systems with up to 37 independent SQUID channels arranged in a hexagonal lattice have been built recently by BTi and Siemens, reducing data-acquisition times for some experiments from hours to minutes. Many experiments have been performed to measure the magnetic fields associated with evoked responses of the brain (see [4] for recent papers). This procedure involves stimulating the subject with a brief audio, visual, or somatic impulse, and observing the magnetic fields produced by the brain in response to the impulse. The idea is to observe these fields at many different spatial locations near the head over several hundred milliseconds after the impulse, with a temporal resolution on the order of milliseconds. There are many variations on the types of impulses that can be presented and whether or not the subject is asked to perform cognitive tasks based upon the type of impulse. By observing the magnetic fields and estimating current sources from them an idea of the brain structures that are participating in the evoked response can be developed. Thus, the evoked response in biomagnetism is a very powerful tool for studying spatio-temporal relationships of the brain, and it will form the model of the kind of data from which the coherence maps discussed below will be derived. A further advantage of the evoked response is that it is repeatable: thus the time sequences of many different evoked responses are in practice averaged to give a good signal-to-noise ratio (SNR) in the measured magnetic fields. Once the magnetic data has been measured over the evoked response, it can be mathematically correlated with itself both in space and time. Knowing the physical model mapping the unknown current sources to the magnetic measurements, the spatio-temporal correlation map of the sources, referred to in this paper as the “biocoherence” of the source, can then be estimated from the correlation of the measurements. Such a map would show for any source point at location r l the relative probability that any other source point at r, is positively active, negatively active, or inactive at a time delay T in relation to the point at r l , averaged over the evoked response. The biocoherence is a natural specialization of the well-known theory of mutual coherence in optics [15], [16] to quasi-static fields. The quasi-static approximation is a good approxi- mation in biology, and implies three constraints that are satisfied by the system [17]: capacitive effects are negligible (i.e., the system is purely resistive), inductive effects are negligible (i.e., the electric field is due solely to the gradient of the electric potential), and propagation delays can be ignored. With the biocoherence map it would be possible to automatically search for strong positive or negative correlations at all spatial separations and temporal delays in a particular evoked response, and observe whether these correlations are similar for normal subjects and whether they show evidence of abnormal function in subjects exhibiting pathology. Correlations of structures at large delay times may indicate previously unrecognized causeand-effect relationships between the structures. As is shown later, the biocoherence map at zero delay time can be used as prior knowledge for a linear estimation reconstruction of the spatial current density at a given time. An important direction in electroencephalography related to the theme of this paper has been to look at the correlation of voltage signals from electrodes placed at various parts of the scalp during different observer tasks [ 181, [ 191. This paper extends some of these ideas as follows: 1) a source-to-data model, utilizing the physical principles of the magnetic-data formation in the quasistatic case, is used to relate the correlations of the magnetic data directly to the source correlations, 2) a method is presented to reconstruct the source correlations from the data correlations, and 3) the estimated source-correlation matrix at zero delay time is used to form reconstructions of the current sources. RELATIONOF THE MAGNETICFIELDTO THE CURRENTDENSITY The Biot-Savart law [20] relates the magnetic induction B(r, t ) (referred to subsequently as the magnetic field) to the current density J(r, t ) : where r is the position of the mgnetic field B ( r , t ) occurring at time t , r’ is the position of the current density J ( r ’ , t ) occurring at time t , and po is the permeability of free space. The position vector r is considered to be outside the source volume in this development, so that B(r, t ) can be noninvasively measured. Bold quantities will indicate vectors or tensors. If the current sources are in a conducting medium, the total current density J ( r ’ , t ) will contain two parts: + J(r‘, t ) = Jp(r’,t ) u(r’, t ) E ( r f ,t) (2) where J p ( r r ,t ) is the primary source current, u ( r r , t ) is the conductivity of the current-containing medium (assumed to be linear and isotropic), and E ( r ‘ , t ) is the electric field induced in the conducting medium by the primary currents. The quantity oE is often referred to as the return current. This current is necessary to avoid build up of charges due to the primary current. SMITH: ESTIMATION OF SPATIO-TEMPORAL CORRELATIONS OF BIOLOGICAL ELECTRICAL SOURCES If the conducting medium is made of a series of N homogeneous conducting regions ok, each region separated by the N - 1 surfaces Sk, then (1) can be written as [22]: N- I B(r, t ) = 999 tion will be referred to as biocoherence. The biocoherence tensor is first defined for the magnetic measurements, and then it is linked to the biocoherence tensor of the source. The Cartesian components of the biocoherence tensor rBrln(7) of the measured magnetic field can be defined as the following ensemble average: + 7) - [rBrln(7)l,/j’ (Ba(r1, t + 7 ) >) - Bp(r2, 0 - (Bp(r2, 0 ) ) ) where Bp(r,t ) = r - rr Jp(rr,t ) x ) r - r r I 3d 3 r ’ , 4~ source - V ( r ‘ , t ) is the electric potential on the surface of interest = -VV(r’, t ) ) ,and Ak(r’)is the outer normal to the kth surface separating the regions of homogeneous conductivity ak from a k + I . (E(rr,t ) BIOCOHERENCE The relationship between the source currents and the magnetic measurements reviewed above constitute the forward problem. Usually the inverse problem is to be solved, which involves estimating the source currents from the magnetic fields. The biomagnetic inverse problem is ill-posed because there are essentially an infinite number of current sources that produce no external magnetic fields. Because of this nonuniqueness of the current distribution it is necessary to impose prior knowledge on the reconstruction. Methods for estimating the source currents at a fixed time from the measured magnetic fields at (5) where the tildes indicate that the magnetic fields B(r, t ) are random processes, a distinction that has not been made until now. The brackets indicate an ensemble averge over the time t of the biological response. Specifically, where a discrete time sampling of integer length T is assumed for a one-dimensional random scalar f ( x , t ) , the temporal ensemble average is defined as 1 c T-7 (”%, t + 7 ) f ( Y y , 0 ) = T - T k = I f(x, k +M Y , 4. (6) Greek subscripts will refer to the x, y, and z components represents the of vectors or tensors. The tensor rErln(7) correlation of the vector magnetic field at position rl with the vector magnetic field at position r2 at a delayed time 7 . This correlation can be obtained computationally after all of the magnetic data has been collected at all measurement positions over the time of the biological response. This is the tensor form of the mutual coherence function that was first defined by Wolf for optical fields [15]. It will be assumed that all random processes are zero mean in the following development to simplify the expressions. Equation (5) then becomes (7) the corresponding time and the types of prior knowledge used in these estimates have been reported extensively in the literature [7]-[ 141. Instead of forming images of the sources at fixed times the quantitative spatio-temporal coherence, or correlation, between the sources that arises solely from their biological behavior can be defined. A direct approach is to determine the spatio-temporal coherence of the magnetic measurements, and from this estimate the coherence of the source, without the need to form intermediate images of the source. That is the approach taken here. The coherence of the sources due solely to their biological func- where (1) has been used, and the primary and return currents have been lumped together into J to simplify the development. The labels s ’ and s r r both represent the source volume. Expressions similar to that above can be defined to describe the coherence of various spatial derivatives of the magnetic field as well. It is useful at this point to write down the expression for the cross product of two vectors in terms of the LeviCivita symbol E [23] so that (8) can be developed further: [ A X Bl, = cP r eolPyAaBy (9) where €123 = €231 = €312 = 1, € 1 3 2 = €213 = €321 = - 1, with all other permutations = 0. Thus (8) can be written as IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. VOL. 39, NO. 10. OCTOBER 1992 loo0 where [rjr,,,(7)lYp E ( l y ( r f ,t + 7)l , , ( r f f ,4 ) (13) the exact counterpart of the magnetic coherence function of (7). Equation (12) links the coherence of the sources with the coherence of their mgnetic fields. Note that rJr,r. ( 7 ) includes the biocoherence of both the primary and return currents. Because the return currents depend linearly upon the primary currents, the biocoherence of the primary currents alone can be extracted. The scalar and vector biocoherences can be directly derived from the tensor biocoherence. This is demonstrated for the magnetic fields: where “ ” and “ X ” represent the vector dot and cross products, respectively. DISCRETEFORMULATION To implement the calculation of the magnetic biocoherence from actual measurements, and to develop estimates of the source biocoherence, it is necessary to discretize the entire problem. In doing so, each Cartesian component of the continuous magnetic data can be approximated by an expansion over a finite set of orthonormal basis functions &(r),i = 1 * * * M , defined in the measurement space (see Appendix A of [14]): M Ba(r, t ) = C B,i(t) d)i(r)? (16) 1 with a similar expansion for each Cartesian component of the continuous source distribution in terms of a finite set N , deof orthonormal basis functions $ i ( r ) ,i = 1 fined in the source space: - t) = 6.. rJ = IajCt) Ic.j(r). (17) Note that these orthonormal basis sets may not be complete, and the expansions represented in (16) and (1 7) will in general provide us with a least squares approximation to the quantity of interest. It is also necessary to point out that the 4i (r)basis set is not related to the sensitivity functions (i.e., lead fields) of the mgnetic detectors. The quantities to be associated with the detector lead fields ’j +i(r) 4jjcr) d3r9 (6ii = 1 when i = j , it is zero otherwise) the Bai(t)are given by they are describing are random processes. Linking this discussion with experiment, the +i(r) could represent the spatial sensitivity of a SQUID coil at location i to the magnetic field at the SQUID coil, so that Bai( t )would be the discrete temporal signal from that particular coil for the a t h component of the field. Again, the ~ $ ~ are ( r )not the lead fields of the system. The lead fields link the sensitivity of a detector to the source elements, and are not generally orthonormal. In contrast, the &(r) basis functions map the continuous magnetic field in a particular region near a SQUID coil to a discrete detector output from that coil. The 4i(r) are independent of the source model: they are sampling functions. The are orthogonal to each other as long as the SQUID coils do not overlap. To connect the expansion coefficients of the magnetic fields to the expansion coefficients of the sources, it is first necessary to write down the continuous linear expression relating one vector field B(r) to another vector field J ( r ) : BJr) N la(r9 will be discussed below, when the full matrix formalism is introduced. The time-dependent expansion coefficient Bai(1) is found as follows. Multiplying the left and right sides of (16) with 4j(r),integrating both sides of the equation over the domain of 4j ( r ) ,and using the orthonormality condition = 5 P s s’ Qap(r, r ’ )J p ( r ’ ) d 3 r f (20) where Qap(r,r f) is the ath, Pth component of the 3 x 3 shift-variant tensor Q(r,r ’ ) relating the two fields, and s ’ is the region over which Ja(r’)is non-zero. Comparing the form of (20) with (3), Q(r,r ‘ ) can be written as the sum of two terms: the first representing the simple vector product associated with the primary current, and the second representing the conduction-dependent retum current, which also depends linearly upon the primary cur- SMITH: ESTIMATION OF SPATIO-TEMPORAL CORRELATIONS OF BIOLOGICAL ELECTRICAL SOURCES rent. Analytic expressions for this second term will in general be complicated, even for simple conducting boundaries. Because of this complexity, empirical mapping of n(r, r f) would be more practical and realistic than its analytic determination. This is discussed briefly below. If (17) is substituted into (20), and this result then substituted into (18), the following expression relating the source and magnetic-field coefficients emerges: ~ a i ( t )= 3 N B j c j’ j’ detector source +i(r)Qafi(rT r ’ ) $ , ( r f )d 3 r rd3r Jfij(t). (21) By introducing the compound indices m = a + 3(i - 1) and n = p 3 ( j - l), the above expression can be written in a more compact form as + 3N gm(t) = (22) where g m ( t ) = Bai(t),L(t)= JBj(t), and Am = = 4 + 3 ( i - 1 ) fi+3(j-1) 1 +;(r)QaP(r, r f ) $ j ( r ’d)3 r fd3r. detector source (23) The elements of the 3M X 1 vector g(t) and the 3N X 1 vector f (t) represent a particular ordering of the expansion coefficients of the magnetic fields and the source fields, respectively, which is defined by the compound indexes m and n. For the case of the data vector this ordering takes the form: g%>= [B,,(t), BYl(t),Bzl(~)7 Bx2(t),By2(~), Bz2(t), * - is general enough to include the effects of the conducting boundaries, assuming that the conduction current is a linear function of the primary current. The A matrix can be found empirically from phantom studies by mapping out the magnetic fields produced by a small current dipole situated at sequentially different locations inside a conducting volume. The conducting volume need not be homogeneous: the effect on the return currents of any realizable conductivity distribution can be built into the A matrix. The formalism is quite powerful in this regard; the difficult part is actually finding the elements of A for a particular problem. At this point the connection between the discrete coherence and the continuous coherence can be made. The covariance matrix (which could also be called the coherence matrix in the present context) of the noisy, discretized measurements can be defined in the following way: Kg(T) AmnSn(t) B,,@), B,,(t), (24) with a similar expressin forfT(t). The “T” superscript indicates the transpose operation. The g ( t ) vector is an ordered list of the SQUID measurements. Equation (22) can be written even more succinctly as g(t) = Af(t) + ii(t) (25) where A is a 3M X 3N array with elements Amn.The zeromean noise vector Z ( t ) , ordered exactly as the data vector g ( t ) , has been introduced to represent additive noise processes unavoidably present in the detection operation. This noise will be correlated and nonstationary in general. Physically, each row i of the A matrix represents the lead field for the corresponding ith detector. The lead field is the sensitivity of a given detector element to the entire source distribution in the presence of conducting boundaries. Each columnj of the A matrix represents the “pointspread function” (PSF) due to a given source element$(t) in the presence of conducting boundaries. The PSF (also referred to as a Green’s function) is the response of all of the detectors due to a single source element. The A matrix 1001 = (g(t +” 7 (26) where it is assumed as before that all random quantities are zero mean, and again the brackets indicate an ensemble average over time. The matrix K,(7) is a 3M X 3M matrix. Using (1 8) and (26) and the definitions below (22), the relationship between the m, nth element of the discrete covariance matrix of (26) and the a, 0th element of the continuous biocoherence tensor r B r r ’ (7)is given by [ K g ( ~ ) I m n= ( g m ( t + 7)Sn(t)) (27) where the indexes m, n are related to CY, 0 and i, j as discussed under (21). Note that noise correlations are implicitly contained within r B r r ’ ( T ) , because noise will be present in the magnetic data. As a specific example of (27) assume that the basis functions represent shifted delta functions: 4;(r) = 6(r - r ; ) , 4 j ( r ’ )= 6(rf - 5 ) (28) so that (27) becomes: rKg (7)Jmn = IrBr,q (7)14- (29) In a real experiment the covariance matrix of (26) will be formed directly from the discrete data. Equation (27) justifies our interpretation of this matrix as the discrete representation of the tensor I’Brrr (7)for a given choice of measurement functions +;(r). Using (25) and (26) it is straightforward to show that, for noise independent of the object (without restricting ourselves to stationary noise or stationary object ensembles), the covariance matrix of the noisy data can be written as: ~ ~ (=7 A) K ~ ( ~ + ) K,(T) A~ (30) where the covariance matrix of the source is defined as Kf(7)= ( f ( t +~)f(t)~>, (31) 1002 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 39, NO. 10, OCTOBER 1992 a 3N is X 3N matrix, and the covariance matrix of the noise &(7) = (ii(t + 7)ii(QT) (32) a 3M X 3M matrix. Equation (30) is the desired result of this section, namely a link between the source and data coherences, and it is the discrete equivalent of (12) with the explicit addition of noise. The A matrix contains the physical modeling of both the primary current and returncurrent contributions to the magnetic fields. The source covariance matrix Kf(7)represents the coherence of the pirmary currents alone, because all of the return-current dependencies are lumped into the A matrix. This is possible because the return currents are linearly dependent upon the primary currents given our assumption of a linear conductor. The relationship between the source covariance matrix Kf(7)and the biocoherence tensor of the sources [see (13)] is analogous to the expression in ( 2 7 ) , with s and s’ both representing the source volume: [Kf(~)~mn ( fm(t p:. The SVD representation of the Q pseudoinverse A+ of A is X P Moore-Penrose R A+ e 1 c-viyT i Pi (36) where the p i are nonzero for i = 1 * R . In practice, the pseudoinverse of (36) must be regularized to account for noise amplification and limitations in numerical precision for small singular values p i . One way this regularization can be implemented is by replacing the coefficient l / p i with p i / @ ? + E ) , and take the limit as E goes to zero. A value of E can usually be found to optimize the performance of the pseudoinverse. A method for modifying these coefficients to include statistical prior knowledge about the object and noise has been demonstrated [ 2 5 ] , [261* Using ( 3 0 ) , the MP estimate of the matrix Kf(7)ATis Taking the transpose of both sides, + 7)JI,(l)) again operating from the left with A + , the M P solution for Kf(7)is obtained: (33) K;.(T)= A+(K,(7) - Kn(7))AfT (39) where the indicates that the quantity is an estimate. The ESTIMATION OF THE SOURCE COHERENCE FROM THE Kg(7) covariance is derived directly from the magnetic data MAGNETIC COHERENCE [cf. ( 2 6 ) ] . The Kn(7) covariance is derived directly from The estimation of the source coherence Kf(7),like the noise measurements in the absence of signal. The absence estimation of the source itself, is an ill-posed inverse of signal could mean that the patient is not present, or that problem in biomagnetism. This ill-posedness is due to the the patient is present, but is not being stimulated to form nature of the Biot-Savart operator relating the sources to evoked responses. The A matrix can be found either from the fields. It is thus necessary to impose prior knowledge mathematical modeling, or as stated earlier from phantom on the quantity being estimated to regularize the solution. studies by mapping out the magnetic fields produced by a The approach to estimating K’(T)here will be to find current dipole situated at sequentially different locations the minimum-norm, least squares solution, also referred inside a conducting volume. The important point is that to as the Moore-Penrose (MP) pseudoinverse [ 2 4 ] . This the expressions on the right-haFd side of the equation can choice is made primarily because of its simplicity, and be found explicitly, allowing Kf(7)to be estimated. Note because of the lack of knowledge of the properties of Kf(7) that 447) may not represent the complete tensor rBr,,J7) in general. The singular value decomposition (SVD) [24] because not all of the Cartesian components of the magof the A matrix will be used to implement this pseudoin- netic field are measured independently at each *measureverse. ment position with standard SQUID techniques. Kf(7)may If a P X Q matrix A is of rank R , it can be expanded still represent the full source tensor rJrjr+), however, if in terms of R unit-rank P X Q outer-product matrices the A matrix is constycted generally enough. formed by the vectors yi and vi: The source matrix KAT) is in essence forming a highly parametrized “signature” for a given patient doing a speR cific neural task. This matrix represents a seven-dimenA =C piyiq‘ (34) sional data set for a three-dimensional source. The phys1 ical significance of the source matrix can be described as where follows. Given any three-dimensional source location of ATAvi = p?vi and AA’y, = p:yi. (35) interest and a desired delay time 7 , the three-dimensional spatial map showing the correlation of all other sources The Q X 1 vectors vi,i = 1 * * Q, and the P X 1 vectors with the source of interest at that delay time averaged over yi,i=l P , are eigenfunctions of A*A and AAT, the evoked response is obtained from the biocoherence respectively, with eigenvalues (called singular values) matrix. This map can give information about which struc- - 0 . . SMITH: ESTIMATION OF SPATIO-TEMPORAL CORRELATIONS OF BIOLOGICAL ELECTRICAL SOURCES tures are linked directly or indirectly, on average, to the source point of interest, and whether this linkage is excitatory or inhibitory. Particularly strong positive or negative correlations can be automatically found from the matrix, with their associated delay times, and used as reference points for a given evoked response by a given subject. Comparison of biocoherences between normal subjects performing the same tasks would allow common features to be determined. Significant departures from these features might indicate abnormal function. Once the estimate of the covariance matrix of the sources has been obtained, it can be used, at zero delay time 7, to produce an estimate of the source distribution at any time t for which the data has been collected: This expression is often referred to as a generalized Wiener filter, and has been demonstrated as a way in which non-stationary sources imaged with shift-variant operators (of which the Biot-Savart law in the presence of conductors is an example) can be estimated [ 141, [25]. This filter minimizes the mean square error between the source and the reconstructed source averaged over the source and noise ensembles. Because of the direct estimation of the source covariant: from the magnetic covariance in (39), reconstructions f ( t ) will be obtained that automatically satisfy the measured magnetic coherences. SUMMARY The goal of this paper was to present a quantitative, formally based framework for unifying the spatial and temporal aspects of biological current sources. The relation between electric-current sources and the magnetic fields that they produce was reviewed. Upon this framework the biocoherence tensor was defined, based upon the mutual coherence function of optics. The term biocoherence was introduced to represent the quasi-static form of the mutual coherence, suitable for the case of biological current sources. A practical discretization procedure was applied to the continuous problem. In the framework of the discretized form, the Moore-Penrose estimate of the source biocoherence was formed from the magnetic biocoherence, which can be directly determined from the magnetic measurements. This source biocoherence may contain clinical and cognitive information regarding the performance of the nervous system over an evoked response. It was then demonstrated how the source biocoherence can itself be used as prior knowledge in the optimum linear estimator to form a reconstruction of current sources at fixed times during the evoked response. An effort is underway for practical applications of the theoretical framework presented in this paper. There was no discussion here of any prior knowledge relating to the structure of the source and field coherences. Further investigation will be made of possible constraints that the coherence matrices may be subject to given the effect of 1003 the conducting medium on the sources’ magnetic-field vectors (e.g., the external magnetic fields due to radial current components within a spherical conductor vanish [22]) and the fact that the divergence of the total current density must be zero. An example of a biocoherence tensor derived from actual SQUID data is unfortunately lacking here; this demonstration will be pursued with multichannel SQUID data when available. REFERENCES [ 11 L. A. Geddes, Electrodes and Measurement of Bioelecrric Events. New York: Wiley-Interscience, 1972. 121 S . J. Williamson, G.-L. Romani, L. Kaufman and I. Modena, Eds., Biomagnefism, An Interdisciplinary Approach. New York: Plenum, 1983. [3] S. J. Williamson and L. Kaufman, “Application of SQUID sensors to the investigation of neural activity in the human brain,” IEEE Trans. M a g . , vol. MAG-19, pp. 835-844, 1983. [4] S. J. Williamson, M. Hoke, G . Stroink and M. Kotani, Eds., Advances in Biomagnetism. New York: Plenum, 1989. 151 J . E. Zimmerman, P. Thiene, and J. T . Harding, “Design and operation of stable rf-biased superconducting point-contact quantum devices, and a note on the properties of perfectly clean metal contacts,” J . Appl. Phys., vol. 41, pp, 1572-1580, 1970. [6] B. N. Cuffin and D. Cohen, “Comparison of the magnetoencephalogram and electroencephalogram,” Elecfroencephal. Clin. Neurophysiol., vol. 47, pp. 132-146, 1979. [7] M. Singh, D. Doria, V . W. Henderson, G. C. Huth and J. Beatty. “Reconstruction of images from neuromagnetic fields,” IEEE Trans. Nucl. Sci., vol. NS-31, pp. 585-589, 1984. [8] W. J. Dallas, “Fourier space solution to the magnetostatic imaging problem,” Appl. Opt., vol. 24, pp. 4543-4546, 1985. (91 R. J. Ilmoniemi, M. S. Hamalainen and J. Knuutila, “The Forward and inverse problems in the spherical model,” in Biomagnetism: Applications and Theory, H. Weinberg, G. Stroink and T. Katila, Eds. New York: Pergamon, 1985, pp. 278-282. IO] W . E. Smith, W. J . Dallas, H . A. Schlitt and W. H. Kullmann, ”Reconstructing a vector current distribution from its magnetic field using linear estimation theory,’’ presented at Top. Meer. Signal Recovery and Synthesis I I , Honolulu, HI, 1986. I l l W. J. Dallas, W. E. Smith, H. A. Schlitt and W. H. Kullmann, “Bioelectric current image reconstruction from measurements of the generated magnetic fields,’’ Medical Imaging, E. H . Schneider and S . J. I . Dywer, Eds., Proc. SPIE, 767, pp. 2-10, 1987. 121 B. Jeffs, R. Leahy, and M . Singh, “An evaluation of methods for neuromagnetic image reconstruction,” IEEE Trans. Biomed. Eng., vol. BME-34, pp. 713-723, 1987. 131 W . Kullmann and W. J. Dallas, “Fourier imaging of electrical currents in the human brain from their magnetic fields,” IEEE Trans. Biomed. Eng., vol BME-34, pp. 837-842, 1987. 141 W . E. Smith, W. J . Dallas, W. H . Kullmann, and H. A. Schlitt, “Linear estimation theory applied to the reconstruction of a 3-D vector current distribution,” Appl. O p t . , vol. 29, pp. 658-667, 1990. 1151 E. Wolf and L. Mandel, “Coherence properties of optical fields,’’ Rev. Modern Phys., vol. 37, no. 2, pp. 231-287, 1965. (161 J. W. Goodman, Srafistical Optics. New York: Wiley, 1985. [ 171 R. Plonsey, Bioelectric Phenomena. New York: McGraw-Hill, 1969. (181 A. S . Gevins, J. C . Doyle, B. A. Cutillo, R. E. Schaffer, R. S. Tannehill, J. H. Ghannam, V . A. Gilcrease, and C . L. Yeager, “Electrical potentials in human brain during cognition: New method reveals dynamic patterns of correlation,” Science, vol. 213, pp. 918-922, 1981. [ 191 A. S . Gevins, J. C . Doyle, B. A. Cutillo, R. E. Schaffer, R. S. Tannehill, and S. L. Bressler, “Neurocognitive pattern analysis of a visuospatial task: Rapidly-shifting foci of evoked correlations between electrodes,” Psychophysiol., vol. 22, no. I , pp. 32-43, 1985. [20] J . D. Jackson, Classical Electrodynamics. New York: Wiley, 1975. [21] D. B. Geselowitz, “On the magnetic field generated outside an inhomogeneous volume conductor by internal current sources,” IEEE Trans. M a g . , vol. MAG-6, pp. 346-347, 1970. 1004 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 39, NO. IO, OCTOBER 1992 [22] J. Sarvas, “Basic mathematical and electromagnetic concepts of the biomagnetic inverse problem,” Phys. Med. Biol., vol 32, pp 1122, 1987. 1231 G. Arfken, Mathematical Methods for Physicists. New York: Academic, 1970. 1241 W K. Pratt, Digital Image Processing. New York: Wiley, 1978 [25] W . E. Smith and H. H . Barrett, “Linear estimation theory applied to the evaluation of a priori information and system optimization in coded-aperture imaging,” J Opt. Soc. Ameri A , vol. 5, pp. 315330, 1988. [26] R. G. Paxman, H. H Barrett, W. E Smith, and T. D. Milster, “Image reconstruction from coded data: 11. Code design,” J. Opt Soc. Amer A , vol. 2 , pp 501-509, 1985. Warren E. Smith (M’88) received the Ph.D. degree in optical sciences from the University of Arizona, Tucson, in 1985. He then visited Germany as a Humboldt Fellow and joined the faculty of the Institute of Optics at the University of Rochester in 1988 His research interests span a wide range of topics in medical imaging He has published in the areas of biomagnetism, automated chromosome analysis, quantitative endoscopy, magnetic resonance imaging, and nuclear medicine. He is also active in the application of fluorescent confocal microscopy to the study of biological and nonbiological objects.