Brivo CT315 Brivo CT325 INSTALLATION GE HEALTHCARE REV 7 3-3 5478821-1EN FIX TABLE AND GANTRY INSTALLATION(CONTINUED) NOTICE If any obstruction is found under the proper anchoring position, stop drilling and use the alternate position. Use the alternate positions only in the case where its corresponding proper position is not possible. (See Illustration 3-11) Gantry Anchor Positions (Proper and Alternative) Positioning Subsystem Illustration 3-11 d. Use the vacuum cleaner to suck up the concrete dust from the hole. Also vacuum dust from the surrounding area when finished. e. Hammer the anchor bolt into the hole made in step b. Illustration 3-12 Anchor Bolt 4. Locate the four leveling pads, and let each of anchor bolts go through the leveling pad respectively. 5. Position the Gantry over the anchor bolts appropriately. 6. Lower the Gantry to close to the floor and position each of leveling pads beneath its associated adjuster. 3-11 Brivo CT315 Brivo CT325 INSTALLATION GE HEALTHCARE REV 7 3-3-2 7. 5478821-1EN Gantry Level Adjustment Place the bubble level at the positions shown as Illustration 3-13. Level Specification: +/- 0.2 degree Illustration 3-13 8. Gantry Adjuster and Bubble Level Position Level the Gantry using the Gantry/Table Anchor Adjuster JIG.(5389135, shipped with system shipping collector). NOTICE The distance between the Gantry stationary base and floor is 5~15mm. 3-12 Brivo CT315 Brivo CT325 INSTALLATION GE HEALTHCARE REV 7 3-3-3 9. 5478821-1EN Gantry Anchoring Set the anchor washer (2105873, shipped with system shipping collector), then tighten the anchor nut to the specified torque to fix the anchor bolts into the floor. Torque: 50 Nm NOTICE If specified torque cannot be achieved, another anchor point must be used. See Illustration 3-11 for alternative anchor position. 10. Loosen all of the anchor nuts, then re-tighten the anchor nuts securely (less than 50 Nm). NOTICE Illustration 3-14 Positioning Subsystem Do not tighten the anchor nuts to the specified torque (50 Nm), because the anchor bolts, tighten in previous step, has been fixed securely into the floor. So, to over-tighten nuts will damage the frame of the Gantry and also bring an out-of-level Gantry. Gantry Anchor Assembly 3-13