Thermo Scientific FOCUS™ GC Gas Chromatograph Service Manual PN 317 094 95, Revision April 2009 FOCUS™ GC Service Manual April 2009 Edition Part Number 317 094 95 © 2007 Thermo Fisher Scientific Inc. All rights reserved. Printed in Italy Published by Thermo Fisher Scientific S.p.A., Strada Rivoltana, 20090 Rodano - Milan - Italy Tel: +39 02 95059373 Fax: +39 02 95059388 Printing History: First Edition, released February 2003 Second Edition, released July 2007 Third Edition, released April 2009 Disclaimer Technical Information contained in this publication is for reference purposes only and is subject to change without notice. Every effort has been made to supply complete and accurate information; however, Thermo Fisher Scientific assumes no responsibility and will not be liable for any errors, omissions, damage, or loss that might result from any use of this manual or the information contained therein (even if this information is properly followed and problems still arise). This publication is not part of the Agreement of Sale between Thermo Fisher Scientific and the purchaser of a Thermo Fisher Scientific system. In the event of any conflict between the provisions of this document and those contained in Thermo Fisher Scientific Terms and Conditions, the provisions of the Terms and Conditions shall govern. Reference to System Configurations and Specifications supercede all previous information and are subject to change without notice. Trademarks FOCUS™ GC is a trademark of Thermo Fisher Scientific Inc., and its subsidiaries. Other brand and product names may be trademarks or registered trademarks of their respective companies. Declaration Manufacturer: Thermo Fisher Scientific S.p.A. Thermo Fisher Scientific S.p.A. is the manufacturer of the instrument described in this manual and, as such, is responsible for the instrument safety, reliability and performance only if: • installation • re-calibration • changes and repairs have been carried out by authorized personnel and if: • the local installation complies with local law regulations • the instrument is used according to the instructions provided and if its operation is only entrusted to qualified trained personnel Thermo Fisher Scientific S.p.A. is not liable for any damages derived from the non-compliance with the aforementioned recommendations. Thermo Fisher Scientific S.p.A. Strada Rivoltana 20090 RODANO (MI) ITALY Part of Thermo Fisher Scientific Thermo Scientific FOCUS GC - Service Manual 3 Contents About This Manual..................................................................................................................................9 Overview .................................................................................................................................... 9 Conventions Used in This Manual........................................................................................... 10 Instrument Markings and Symbols .......................................................................................... 12 References ................................................................................................................................ 14 FOCUS™ GC Document Set ..................................................................................... 14 Additional References................................................................................................. 14 Classification of the Instrument ............................................................................................... 15 Technical Features ................................................................................................................... 15 Safety Information ................................................................................................................... 16 Use of Gases................................................................................................................ 16 Using Hydrogen ....................................................................................................................... 17 Using the Hydrogen Sensor ........................................................................................ 18 Cleaning and Decontamination ................................................................................................ 18 Chapter 1 Overview............................................................................................................................................. 19 Introduction .............................................................................................................................. 20 How to control the FOCUS GC .................................................................................. 21 The GC Base Unit ....................................................................................................... 21 Instrument Automation ............................................................................................... 22 Internal Automation ....................................................................................... 22 Communication with External Units ............................................................. 22 Autosampler Interface.................................................................................... 22 Data Systems Interface .................................................................................. 22 General Requirements.............................................................................................................. 23 Power Quality ............................................................................................................. 23 Minimum Power Requirements .................................................................................. 23 Gas and Plumbing Requirements................................................................................ 24 Gases Purity Requirements ......................................................................................... 24 Gas Regulators, Gas Lines and Fittings Requirements............................................... 25 Gas Regulators ............................................................................................... 25 Gas Lines ....................................................................................................... 25 Fittings ........................................................................................................... 25 Chapter 2 Additional Technical Information ..................................................................................................... 27 FOCUS GC Overview.............................................................................................................. 28 Thermo Scientific FOCUS GC - Service Manual v Contents FOCUS GC Internal View.......................................................................................... 28 GC Oven ..................................................................................................................... 28 S/SL Injector............................................................................................................... 29 PPKD Injector............................................................................................................. 29 FID Detector ............................................................................................................... 29 TCD Detector ............................................................................................................. 29 Preliminary Tests and Regulations .......................................................................................... 30 Autosamplers Interface............................................................................................................ 38 Kits/Devices Option................................................................................................................. 38 Oven Hydrogen Sensor............................................................................................... 38 Merlin Valve Kit......................................................................................................... 38 Kit LAN for FOCUS GC............................................................................................ 39 Packed Column Installation........................................................................................ 39 Chapter 3 Parts Replacement .............................................................................................................................41 Mechanical Parts Replacement................................................................................................ 42 Oven Section............................................................................................................... 42 Air Flap System .......................................................................................................... 54 Air Flap Section............................................................................................. 54 Pneumatic Modules Section ....................................................................................... 59 Injector Section........................................................................................................... 65 Detector Section ......................................................................................................... 70 Electronic Parts Replacements ................................................................................................ 80 Introduction ................................................................................................................ 80 GC with FID .................................................................................................. 80 GC with TCD ................................................................................................ 81 Electronic Cassette Section ........................................................................................ 81 Electronic Boards Section .......................................................................................... 86 Chapter 4 Electronic Boards Description..........................................................................................................91 General Description ................................................................................................................. 92 Electronic Main Cassette ......................................................................................................... 94 Main Schematic Diagrams....................................................................................................... 96 Introduction to MY04 Main Board.......................................................................................... 97 Power Supply Notes ................................................................................................... 99 General Notes ............................................................................................................. 99 vi Thermo Scientific FOCUS GC - Service Manual Contents Connectors Dismounting on Main MY04 Board ................................................................... 106 MY04 Main Board Sections Description............................................................................... 108 Schematics and Component Layout.......................................................................... 108 Sheet 1/7 (CPU 1) ..................................................................................................... 108 Sheet 2 (CPU2) ......................................................................................................... 109 Sheet 3 (DPFC1) ....................................................................................................... 110 Sheet 4 (DPFC2) ....................................................................................................... 110 Sheet 5 (Power Supply; Valves; Stepper Drivers) .................................................... 111 Sheet 6 (Oven; Zone Heaters; Drivers)..................................................................... 112 Sheet 7 (Oven; Zone Thermoregulators) .................................................................. 112 ELI201 Electrometric Board (FID)........................................................................................ 113 Schematics and Component Layout.......................................................................... 113 Board Description ..................................................................................................... 113 Sheet 1/2 (SP1; V/F; Power Supply)......................................................................... 114 SE 1467 Sheet 2/2 (Auto zero; Gain; Range) ........................................................... 114 CVA201 Electrometer Pre-amplifier Board (FID) ................................................................ 115 Schematics and Component Layout.......................................................................... 115 Board Description ..................................................................................................... 115 TCDI203 Electrometric Board (TCD) ................................................................................... 116 Schematics and Component Layout.......................................................................... 116 Board Description ..................................................................................................... 116 Sheet 1/3.................................................................................................................... 116 Sheet 2/3.................................................................................................................... 117 Sheet 2/3.................................................................................................................... 117 TCDE203 Board (TCD)......................................................................................................... 117 Schematics and Component Layout.......................................................................... 117 Board Description ..................................................................................................... 117 Sheet 1/4.................................................................................................................... 117 Sheet 2/4.................................................................................................................... 118 Sheet 3/4.................................................................................................................... 118 Sheet 4/4.................................................................................................................... 118 FRTMY100 Board and Keypad ............................................................................................. 118 Schematics and Component Layout.......................................................................... 118 Board Description ..................................................................................................... 118 IFM311 LAN Option ............................................................................................................. 119 Schematics and Component Layout.......................................................................... 119 Thermo Scientific FOCUS GC - Service Manual vii Contents Chapter 5 Maintenance and Troubleshooting Notes ........................................................................................121 Maintenance........................................................................................................................... 122 Suggested Maintenance Schedule ............................................................................ 122 Standard Maintenance Schedule............................................................................... 123 Maintenance References........................................................................................................ 124 Maintaining a S/SL Injector ..................................................................................... 124 Refer to: ....................................................................................................... 124 Maintaining a PPKD Injector ................................................................................... 124 Refer to: ....................................................................................................... 124 Maintaining a FID Detector...................................................................................... 125 Refer to: ....................................................................................................... 125 Maintaining a TCD Detector .................................................................................... 125 Refer to: ....................................................................................................... 125 Analytical Troubleshooting Notes......................................................................................... 126 Technical Troubleshooting Notes.......................................................................................... 130 Error Messages ......................................................................................................... 130 Info/Diagnostics........................................................................................................ 132 Hydrogen Sensor Setting .......................................................................................... 135 Changing the GC Manufacturing Data..................................................................... 135 Chapter 6 Schematic Drawings Summary.........................................................................................................137 Drawings List......................................................................................................................... 138 Glossary .............................................................................................................................................. 139 Index.................................................................................................................................................... 145 viii Thermo Scientific FOCUS GC - Service Manual About This Manual Overview WARNING! This Service Manual must be used with the FOCUS™ GC Operating Manual and FOCUS™ GC Site Preparation and Installation Manual. This manual provides detailed information for installing, servicing, and repairing the FOCUS™ GC. This manual is written primarily for certified service representatives. Chapter 1, Overview, contains a general approach to the FOCUS™ GC reporting for your convenience some information already detailed in the FOCUS™ GC Documents Set. Chapter 2, Additional Technical Information, contains important additional technical information. Chapter 3, Parts Replacement, contains detailed instructions to replace the most important mechanical and electronic parts of the FOCUS™ GC. Chapter 4, Electronic Boards Description, contains a detailed description of the electronics boards of the FOCUS™ GC. Chapter 5, Maintenance and Troubleshooting Notes, contains the guides to maintain the S/SL and PPKD injectors and the FID and TCD detector, then and analytical and technical troubleshooting notes. Chapter 6, Schematic Drawings Summary, contains the list of the diagrams. The Glossary contains definitions of terms used in this manual. This also includes abbreviations, acronyms, metric prefixes, and symbols. The Index contains an alphabetical list of key terms and topics in this guide, including cross references and the corresponding page numbers. Thermo Scientific FOCUS GC - Service Manual 9 About This Manual Conventions Used in This Manual Conventions Used in This Manual The following symbols and typographical conventions are used throughout this manual. Bold Bold text indicates names of windows, dialog boxes, and fields. Italic Italic indicates cross references, first references to important terms defined in the glossary, and special emphasis. Monospace Monospace, or Courier, indicates filenames and filepaths or text the user should enter with the keyboard. Monospace Bold Monospace Bold indicates messages, prompts, or menu titles displayed on the computer screen or on a digital display. » This symbol illustrates menu paths to select, such as File»Open…. KEY NAME Bold, uppercase sans serif font indicates the name of a key on a keyboard or keypad, such as ENTER. CAUTION This symbol alerts you to an action or sequence that, if performed improperly, could damage the instrument. NOTE This symbol alerts you to important information related to the text in the previous paragraph. This symbol alerts you to an action or sequence that, if WARNING! improperly performed, could result in damage to the instrument or possible physical harm to the user. This symbol may be followed by icons indicating special precautions that should be taken to avoid injury. This symbol indicates an electric shock hazard. 10 Thermo Scientific FOCUS GC - Service Manual Conventions Used in This Manual About This Manual This symbol indicates danger from hazardous chemicals. This symbol indicates danger from high temperature surfaces or substances. This symbol indicates a fire hazard. This symbol indicates an explosion hazard. This symbol indicates a toxic hazard. This symbol indicates the presence of flammable materials. This symbol indicates the presence of radioactive material. This symbol indicates an operation or sequence that must not be performed by the user. A Thermo Fisher Scientific authorized Customer Support Engineer must perform this sequence. This symbol indicates all metal objects, such as watches and jewelry, must be taken off. This symbol indicates an eye hazard. Eye protection must be worn. This symbol indicates the user must wear a protective screen when performing the sequence. Thermo Scientific FOCUS GC - Service Manual 11 About This Manual Instrument Markings and Symbols This symbol indicates the user must wear protective shoes when performing the sequence. This symbol indicates the user must wear protective clothing when performing the sequence. This symbol indicates the user must wear gloves when performing the sequence. Instrument Markings and Symbols The following table explains the symbols used on Thermo Fisher Scientific instruments. Only a few of them are used on the FOCUS™ GC gas chromatograph. See the asterisk. Symbol Description Direct Current * Alternating Current Both direct and alternating current Three-phase alternating current 3 Earth (ground) terminal Protective conductor terminal Frame or chassis terminal 12 Thermo Scientific FOCUS GC - Service Manual Instrument Markings and Symbols About This Manual Symbol Description Equipotentiality * On (Supply) * Off (Supply) Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION (Equivalent to Class II of IEC 536) * Instruction manual symbol affixed to product. Indicates that the user must refer to the manual for specific Warning or Caution information to avoid personal injury or damage to the product. Caution, risk of electric shock * Caution, hot surface * Caution, biohazard In-position of a bistable push control Out-position of a bistable push control * Thermo Scientific FOCUS GC - Service Manual Symbol in compliance to the Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) placed on the european market after August, 13, 2005. 13 About This Manual References References FOCUS™ GC Document Set Site Preparation and Installation Manual (PN 317 094 11) This manual and diskette describes how to set up a workspace for the FOCUS™ GC and how to connect the FOCUS™ GC to the gas supplies and peripheral devices. Instruction Manual (PN 317 094 12) This manual provides descriptions of the FOCUS™ GC hardware and software and instructions for their use. Spare Parts Catalog (PN 317 094 13) This catalog contains a list of spare parts for the FOCUS™ GC. Additional References Kit LAN for FOCUS GC (PN 190 506 15) - Installation Guide (PN 317 094 82) This guide provides the instruction for installing the Kit LAN on the FOCUS GC. Merlin Microseal™ Kit for S/SL and PTV Inlet System (PN 190 502 05) Installation Guide (PN 317 093 65) This guide provides the instructions to install the Merlin Microseal™ High Pressure Septum kits on the capillary S/SL and PTV inlet systems. AI 3000/AS 3000 Operating Manual (317 093 92) This manual contains description of the features and components of the AI 3000/ AS 3000 Autosampler. Inside, you will find all of the information for routine operation. TriPlus Operating Manual (PN 317 094 41) This manual contains description of the features and components of the TriPlus AS/HS/SPME autosampler. Inside, you will find all of the information for routine operation of your autosampler. 14 Thermo Scientific FOCUS GC - Service Manual Classification of the Instrument About This Manual Classification of the Instrument Environmental Conditions • Internal use. • Altitude up to 2000 meters. • Temperature from 15 to 35 °C. • Maximum relative humidity between 30% and 85%. • Voltage variations not exceeding +/- 10% of the nominal value. • Transients according to installation categories II. • Degree of pollution according to IEC 664 (3.7.3) 2. Technical Features The following table summarizes the major technical features of the FOCUS™ GC Features Description • External Interface — RS232 Serial Line • Power Supply — 120 Vac; 50/60 Hz; 1600 VA — 230 Vac; 50/60 Hz; 1600 VA • Dimensions — 680 x 640 x 440 mm (h x l x w) • Mass — About 30 kg Thermo Scientific FOCUS GC - Service Manual 15 About This Manual Safety Information Safety Information WARNING! When, for technical reasons, it is necessary to work on instrument parts which may involve an hazard (moving parts, components under voltage, etc.) the authorized Field Technical Service must be contacted. This type of situations can be identified because access to these parts is possible only by using a tool. The removable protective covers bear a warning symbol suggesting to refer to the documentation accompanying the instrument. Should an operator perform a maintenance operation, he/she must have received proper training to carry out that specific action. Before using dangerous substances (toxic, harmful, etc.), please read the hazard indications and information reported in the Safety Sheet supplied by the manufacturer referring to the relevant CAS (Chemical Abstract Service) number. The instrument must be used according to the specifications of this guide. Improper use can adversely affect the instrument protection. The analytical operation requires the use of chemical substances having different hazard specifications. Before using chemicals, please read the hazard indications and information reported in the Safety Sheet supplied by the manufacturer referring to the relevant CAS (Chemical Abstract Service) number. Use of Gases WARNING! Before using gases, carefully read the hazard indications and information reported in the Safety Sheet supplied by the manufacturer referring to the CAS (Chemical Abstract Service) number. It is the user’s responsibility to see that all local safety regulations for the use of gases are obeyed. 16 Thermo Scientific FOCUS GC - Service Manual Using Hydrogen About This Manual Using Hydrogen The use of hydrogen as a carrier gas or as fuel for certain flame detectors requires the operator’s strict attention and compliance with special precautions due to the hazards involved. Hydrogen is a dangerous gas, particularly in an enclosed area when it reaches a concentration corresponding to its lower explosion level (4% in volume). When mixed with air it can create an explosive mixture. An explosion hazard could develop in the GC oven when hydrogen is used as a carrier gas if oven elements are not perfectly connected to each other, or if the connection materials are worn out, broken, or otherwise faulty. Use the following safety precautions when using hydrogen: • Ensure that all hydrogen cylinders comply with the safety requirements for proper use and storage. Hydrogen cylinders and delivery systems must comply with local regulations. • Make sure the gas supply is turned completely off when connecting hydrogen lines. • If an electronic leak test meter should not be available, perform a bubble test to ensure that the hydrogen lines are leak-tight before using the instrument. Perform this test after the pressure test described in the Testing for Leaks paragraph in FOCUS™ GC Site Preparation and Installation Guide Manual. Repeat this test to eliminate all leaks. • Ensure your GC column oven has a Thermo Fisher Scientific hydrogen sensor. A hydrogen sensor continuously monitors the hydrogen level in the GC column oven. If your GC oven does not have a hydrogen sensor already installed, contact your Thermo Fisher Scientific sales representative. To comply with instrument safety requirements, a Thermo Fisher Scientific CSE or authorized service technician should install the sensor. If you plan to use a sensor other than the recommended Thermo Fisher Scientific sensor, you must verify its ability to perform the functions listed above before installing it. It must comply with your local safety regulations, or with the IEC 61010 regulations if local regulations do not exist. Thermo Scientific FOCUS GC - Service Manual 17 About This Manual Cleaning and Decontamination Using the Hydrogen Sensor The lower limit of the hydrogen sensor is 0.5% in volume. You should adjust the detection threshold to 1% in volume, which is 25% of the hydrogen lower limit of explosion (4% in volume). In cases where the connections begin to leak or the column breaks, the sensor alerts the operator. Then it automatically cuts off the gas supply and heating to the active zones, and sweeps the column oven with forced air ventilation. If the sensor detects anomalies or leaks during GC operation due to instrument malfunction, the operator must immediately: • close the hydrogen supply • switch off the gas chromatograph • air out the room The reliability of the sensor depends on careful maintenance. After the sensor is in use, you must periodically check its operating performance and calibration as recommended by the manufacturer. Refer to your hydrogen sensor’s instruction manual for maintenance guidelines. WARNING! Never use hydrogen in FOCUS™ GC unless your GC oven has a hydrogen sensor installed. Thermo Fisher Scientific CSEs are not authorized to install or repair any instrument using hydrogen as a carrier gas unless the instrument is equipped with the appropriate sensor. Cleaning and Decontamination Normal usage of the FOCUS™ GC can cause the exterior to get dirty. Clean the outer surfaces by wiping them with a cloth dampened with water. In the event that a hazardous material is spilled on or in the instrument, clean the spill according to the procedures in the Material Safety Data Sheet for that substance. 18 Thermo Scientific FOCUS GC - Service Manual Overview 1 This chapter contains an overview of the FOCUS™ GC reporting for your convenience some information already detailed in the FOCUS™ GC Documents Set. Recommended test equipment and suggested tools are also included. Chapter at a Glance… Introduction ..........................................................................................................20 General Requirements ..........................................................................................23 Thermo Scientific FOCUS GC - Service Manual 19 Chapter 1 Overview Introduction Introduction The available configurations of the FOCUS GC are: • S/SL injector on the right side and a FID detector on the left side of the GC. • PPKD injector on the right side and a TCD detector on the left side of the GC. The carrier gas to the injector is supplied through a DPFC module, instead the detector gases are supplied to the detector base body through a conventional pneumatic module (non-DGFC). The pneumatic modules are located on the top rear side of the GC. The parts interested are: • S/SL injector • PPKD injector • FID detector • TCD detector • Injector and detector gas control modules • PT100 oven temperature sensor. All the other parts have been new designed for the FOCUS™ GC. NOTE 20 No other configurations and no other detectors type are available. Thermo Scientific FOCUS GC - Service Manual Chapter 1 Overview Introduction How to control the FOCUS GC The GC can be controlled as stand alone or through a dedicated Thermo Fisher Scientific data system. In the first case, all the parameters must be set using the GC keypad and the output detector signal will be processed through a data system such as Chrom-Card, ChromQuest, Xcalibur, etc. In the second case, the data system will control both the GC parameters setting and the output detector signal. The GC Base Unit The GC base unit includes the four major components of the FOCUS GC. • Pneumatic compartment • Analytical unit • Electronic compartment • Status panel and keypad Thermo Scientific FOCUS GC - Service Manual 21 Chapter 1 Overview Introduction Refer to: • Chapter 1 of the FOCUS GC Instruction Manual Instrument Automation FOCUS GC contains several automated features for running the GC, communicating with other analysis equipment, and interacting with a data system. Internal Automation You program internal automation by entering run time events in special menus. You can set these events to execute at specified times after start analysis. Communication with External Units You can connect the GC to external modules and accessories, such as data systems, autosamplers and hydrogen sensor. Autosampler Interface The TriPlus AS, HS, SPME or AI 3000/AS 3000 autosampler can be connected to the GC. Data Systems Interface Your FOCUS GC generates analog and digital data output when you perform chromatographic analysis. A computer with a data system or a computing integrator can be used to process the data from the GC. Refer to: • 22 Chapter 1 of the FOCUS GC Instruction Manual Thermo Scientific FOCUS GC - Service Manual Chapter 1 Overview General Requirements General Requirements This paragraph summarizes general FOCUS GC requirements already described in the FOCUS™ GC Site Preparation and Installation Manual. The GC has been developed to have an easy use but also an easy and fast installation procedure. No particular set must be performed after the GC installation to have the system running. The simple installation without meet inconvenient and troubles will be obtained also if the instructions reported below will be respected. • Power Quality • Minimum Power Requirements • Gas and Plumbing Requirements • Gases Purity Requirements • Gas Regulators, Gas Lines and Fittings Requirements Power Quality The quality of power supplied to your GC system must be stable and within the general specifications listen below. The line must be free of fluctuation due to slow changes in the average voltage, surges sags, transients and harmonics. The GC must be connected according with its power set, to the an acceptable source of power. Minimum Power Requirements 120 Vac 230 Vac Voltage 120 VAC + 6% -10% 230 VAC ±10% Frequency 50/60 Hz 50/60 Hz ± 2 Hz Power requirements 13.3 A 7A Maximum power consumption during full power heat-up 1600 W 1600 W Average power consumption 400 W 400 W Thermo Scientific FOCUS GC - Service Manual 23 Chapter 1 Overview General Requirements Gas and Plumbing Requirements WARNING! All Thermo Fisher Scientific gas chromatographs are designed to use an inert gas as the carrier gas. If you wish to use hydrogen as carrier gas, you must install a hydrogen sensor. Contact a Thermo Fisher Scientific sales representative if you plan to use hydrogen as the carrier gas in your new FOCUS GC. If you do not have the hydrogen sensor, you must use an inert carrier gas. The use of hydrogen gas as carrier gas without the hydrogen sensor installed on the FOCUS™ GC is under the user responsibility only. Gases Purity Requirements The following notes describe the minimum requirements for gases used in your gas chromatograph system: Gas Purity Requirements Helium for carrier gas 99.995% high purity, with less than 1.0 ppm each of water, oxygen and total hydrocarbons after purification. Use a regulator with a CGA 580 connection or equivalent. Hydrogen for detector fuel gas and/or carrier gas 99,995% high purity, with less than 1.0 ppm each of water, oxygen and total hydrocarbons after purification. Use a regulator with a CGA 350 connection or equivalent. Nitrogen for detector fuel gas and/or carrier gas 99.995% high purity with less than 1.0 ppm of total hydrocarbons after purification. Use a regulator with a CGA 580 connection or equivalent. Air for detector fuel gas 99.995% high purity. Use a regulator with a CGA 0590 connection or equivalent. WARNING! Air compressors are not acceptable because they do not meet pressure, water and hydrocarbon requirements. NOTE 24 Use of water, oxygen and hydrocarbon traps are suggested. Thermo Scientific FOCUS GC - Service Manual Chapter 1 Overview General Requirements Gas Regulators, Gas Lines and Fittings Requirements Gas Regulators Gas tanks may be equipped with either single or dual-stage regulators that contain stainless steel diaphragms. The regulator output pressure should be adjustable • from 300 to 1000 kPa (45-145 psi) for the DPFC module • 350 kPa (51 psi) for each line of the non-DGFC module Gas Lines The tubings connected to the FOCUS GC must be new. In case of used tubings, please pay attention that they are oil and moisture free. In case of doubts, they must be cleaned before to be connected to the GC as described in Cleaning and Purging Gas Lines operating sequence in Chapter 2 of this manual. Fittings The GC can be installed by using 2 mm OD and 1/8” OD tubings. According with the tubing used the 2 mm OD and 1/8” OD ferrules must be used. Thermo Scientific FOCUS GC - Service Manual 25 Chapter 1 Overview 26 General Requirements Thermo Scientific FOCUS GC - Service Manual Additional Technical Information 2 This chapter contains additional technical information. Chapter at a Glance… FOCUS GC Overview..........................................................................................28 Preliminary Tests and Regulations .......................................................................30 Autosamplers Interface.........................................................................................38 Kits/Devices Option .............................................................................................38 Operating Sequences Cleaning and Purging Gas Lines ..........................................................................30 External GC Leak Test .........................................................................................31 Injector Manual Leak Test....................................................................................32 Detector Base Body Manual Leak Test ................................................................33 DPFC Module Calibration....................................................................................35 Thermo Scientific FOCUS GC - Service Manual 27 Chapter 2 Additional Technical Information FOCUS GC Overview FOCUS GC Overview This section provides a brief description of the FOCUS GC. FOCUS GC Internal View According to the GC S/SL-FID, S/SL-TCD or PPKD-TCD configuration, the components located inside the GC unit are: • GC Oven • S/SL Injector • PPKD Injector • FID Detector • TCD Detector GC Oven The GC oven, shown in Figure 2-1, includes the following components. • The PT100 oven temperature sensor is separately from the oven diaphragm but positioned and fixed on front of the diaphragm and it could be removed from the top of the GC. • The oven heater (1300 W), located on the rear of the oven diaphragm. • The oven fan located besides the oven diaphragm. 1 1. Oven PT 100 Holder Figure 2-1. Oven Internal View 28 Thermo Scientific FOCUS GC - Service Manual Chapter 2 Additional Technical Information FOCUS GC Overview Refer to: • Chapter 4 of the FOCUS GC Instruction Manual S/SL Injector Split/splitless injector (S/SL) is located on the RIGHT side of the GC. The S/SL injector is gas supplied through a DPFC module (250 or 1000 kPa). Refer to: • Chapter 5 of the FOCUS GC Instruction Manual PPKD Injector Purged packed column injector (PPKD) is located on the RIGHT side of the GC. Refer to: • Chapter 6 of the FOCUS GC Instruction Manual FID Detector Flame Ionization Detector (FID) is located on the LEFT side of the GC. FID is not visible because it is covered from the detector cover. The detector can be reached pulling up its cover manually. FID is gas supplied through a non-DGFC module (350 kPa). Refer to: • Chapter 8 of the FOCUS GC Instruction Manual TCD Detector Thermal Conductivity Detector (TCD) is located on the LEFT side of the GC. TCD is not visible because it is covered from the detector cover. The detector can be reached pulling up its cover manually. TCD is gas supplied through a dedicated DGFC module. Thermo Scientific FOCUS GC - Service Manual 29 Chapter 2 Additional Technical Information Preliminary Tests and Regulations Refer to: • Chapter 9 of the FOCUS GC Instruction Manual Preliminary Tests and Regulations This paragraph contains the instruction to perform the preliminary test and regulation of parts of the FOCUS GC. To perform the operations, refer to the following operating sequences: • Cleaning and Purging Gas Lines on page 30 • External GC Leak Test on page 31. • Injector Manual Leak Test on page 32. • Detector Base Body Manual Leak Test on page 33. • DPFC Module Calibration on page 35. OPERATING SEQUENCES Cleaning and Purging Gas Lines The used tubings must be cleaned before connecting them to the gas regulator and GC operating as follows: 1. Remove all the traps if present. 2. The tubings must be solvent-flushed and purge-dried with an inert gas as helium or nitrogen. 3. Flush gas lines using acetone or hexane, NEVER use chlorinate solvents. 4. When flushing gas lines, collect the solvent passing through the gas line and inspect it for discoloration or residue. Continue to flush gas lines until all waste solvent is free of discoloration and residue. 30 Thermo Scientific FOCUS GC - Service Manual Preliminary Tests and Regulations Chapter 2 Additional Technical Information 5. Purge the gas lines for several minutes with an inert gas having a pressure of 35 kPa (5 psi) to make sure the gas line is free of all traces of solvent. 6. Install the new traps. 7. Connect the cleaned tubings to the GC. OPERATING SEQUENCES External GC Leak Test After the tubings connection to the GC, the external leak test must be performed as follows: 1. Switch OFF the GC. 2. Adjust the air, hydrogen and make up gas regulators to 350 kPa (51 psi). 3. Adjust the carrier gas regulator to 300 kPa (45 psi) or 700 kPa (100 psi) or 1000 kPa (145 psi) according to the pneumatic module installed. 4. Wait some minutes to pressurize well the external gas lines. 5. Close the gas regulators and verify that there is no loss of pressure. WARNING! In case of leaks, do not use the Liquid soap but an electronic leak detector. After the external leak test, the GC installation can continue with the installation of the column (see the FOCUS™ GC Operating Manual). NOTE The internal leak test has been already performed in the factory during the FOCUS GC analytical test. In case of trouble where the injector and detector are suspected of leaks, please follow the instruction reported in the Injector Manual Leak Test and Detector Base Body Manual Leak Test operating sequences. Thermo Scientific FOCUS GC - Service Manual 31 Chapter 2 Additional Technical Information Preliminary Tests and Regulations OPERATING SEQUENCES Injector Manual Leak Test Perform the injector manual leak test operating as follows: 1. Adjust the carrier gas regulator to 300 kPa (45 psi) or 700 kPa (100 psi) or 1000 kPa (145 psi) according to the pneumatic module installed. 2. Close the bottom side of the injector with a cap. 3. Switch ON the GC. 4. After the GC auto test, verify that all the temperatures are set to the room temperature or to 50 °C maximum, then, set the carrier pressure to 100 kPa (15 psi). 5. Close manually the purge and split lines on the injector using the GC keypad. 6. Verify on the display that the ACT pressure value is equal to the SETPOINT value. 7. Wait one-two minutes to pressurize well the injector, then set the carrier gas to OFF. 8. If the system is leak free, the ACT pressure value does not change. WARNING! In case of leaks, do not use the liquid soap but an electronic leak test meter. 32 Thermo Scientific FOCUS GC - Service Manual Preliminary Tests and Regulations Chapter 2 Additional Technical Information OPERATING SEQUENCES Detector Base Body Manual Leak Test NOTE The leal test of the detector base body CANNOT BE performed alone. It is caused from the non-DGFC modules that supply the gases to the detector base body where no pressure or flow is shown on the GC display. For this reason the capillary detector base body leak test MUST BE performed using the carrier gas and connecting the column between the base body and the injector as described on the procedure below: Perform the injector and detector base body leak test as follows: 1. Set the detector gas regulators to OFF. 2. Adjust the carrier gas regulator to 300 kPa (45 psi) or 700 kPa (100 psi) or 1000 kPa (145 psi) according to the pneumatic module installed. 3. Install the test capillary column between the injector and the capillary detector base body. NOTE Instruction on the capillary column installation are provided in the chapter “Columns” of the FOCUS™ GC Operating Manual. 4. Remove the small detector top cover. 5. Only for FID Remove the FID detector loosing the its fixing screw without removing the FID cables from the detector and close the top of the capillary detector base body with a brass cap. 6. Only for TCD Close the carrier/make-up channel outlet with the pressure meter. 7. Switch ON the GC 8. After the FOCUS™ GC auto test, verify that all the GC temperature are set to the room temperature or 50 °C maximum, then set the carrier pressure to 100 kPa (15 psi). The auxiliary detector gases must be set to OFF. Thermo Scientific FOCUS GC - Service Manual 33 Chapter 2 Additional Technical Information Preliminary Tests and Regulations In the case of TCD, the filaments must be OFF. 9. Close manually the Purge and Split lines on the injector using the GC keypad. 10. Verify on the display that the ACT pressure value is equal to the SETPOINT value. Wait five minutes to pressurize well the injector, the test column and the detector base body, then set the carrier gas to OFF. CAUTION If different column length is installed, more time must be required to equilibrate the circuit. 11. If the system is leak free, the ACT pressure value does not change. WARNING! In case of leaks, do not use the liquid soap but an electronic leak test meter. 12. Switch OFF the GC. 13. Only for TCD Remove the pressure meter from the carrier/make-up channel outlet. Close the reference channel outlet with the pressure meter and repeat the operations starting from the point 7. 34 Thermo Scientific FOCUS GC - Service Manual Preliminary Tests and Regulations Chapter 2 Additional Technical Information OPERATING SEQUENCES DPFC Module Calibration Please refer to: • DPFC Module Calibration By Using Kit PN 190 501 20 • DPFC Software Calibration with K=1 DPFC Module Calibration By Using Kit PN 190 501 20 WARNING! This operating sequence must be performed only when required. Only the DPFC carrier gas module could be calibrated by using the kit PN 190 501 20. In addition to the calibration kit, two measurement instruments as Digital Flow meter and Digital Pressure gauge are used to perform the DPFC module calibration. The two measurement instruments are not included in the kit. The J&W ADM1000 Digital Flow meter and the FLUKE Digital Pressure gauge 713 100G are the measurement instruments used in our factory to perform the DPFC module calibration. Different instruments could be used but in this case they must have the following technical characteristics: Digital Flow Meter • Flow range: from 0,1 to 1000 ml/min auto-ranging • Accuracy: ± 3% of reading • Operating temperature range: from 0 °C to 45 °C • Power: 9 V battery (alkaline) • Compatible with all non-corrosive gases Digital Pressure Gauge • Pressure Input range: from 0 to 100 psi (0 - 689,5 kPa) • Accuracy: ± 0,05% of range Thermo Scientific FOCUS GC - Service Manual 35 Chapter 2 Additional Technical Information Preliminary Tests and Regulations • Maximum non-destructive pressure: 2 x top of range (200 psi, 1380 kPa, 13,8 bar) • Operating temperature: from -10 °C to 55 °C Perform the DPFC carrier module calibration as follows: 1. Switch OFF the GC and connect the RS232 cable between the GC and the PC. 2. Load in the PC the calibration software (PN 274 000 20) included in the kit. 3. Replace the current injector septum cap with the relative joint included in the kit. 4. Connect the Kit Calibration Box, the measurement instruments and the injector joint as shown in Figure 2-2. Not Used 2 DETECTOR FLUKE PRESSURE GAUGE FLUKE PRESSURE GAUGE INJECTOR 1 3 A B To Injector J&W FLOW METER J&W FLOW METER Figure 2-2. Kit Calibration Box Connections 5. Remove the column from the injector if present and close its bottom inlet with a cap. 6. Switch ON the GC. 7. After the GC auto test, set the carrier pressure to 30 kPa. 36 Thermo Scientific FOCUS GC - Service Manual Preliminary Tests and Regulations Chapter 2 Additional Technical Information 8. Set the injector and oven temperatures to 40 °C. 9. On the PC, launch the calibration software and follow the instructions shown on the video. 10. When the injector calibration procedure has been completed, power the GC Off. Assemble the injector as the analytical method required, then power ON the GC. DPFC Software Calibration with K=1 This new calibration procedure has been developed mainly to help the Field Service Engineer to perform checking and troubleshooting on the DPFC (Digital Pressure Flow Controller) pneumatic modules in order to avoid an unnecessary replacement of the DPFC module when the calculated K-Factor is far to the nominal one. Before performing this procedure, it is important to check all other possible causes such as leaks, column length, carrier gas type, etc. Tools Required The perform the procedure the following tools are required: NOTE • DPFC Software Calibration Version 3.6.1.The software can be dowloaded free from the Milan Tech Support Web www.ceinstruments.it. • Calibrated Restrictor for DPFC with K=1.the restrictor is equipped by the own Certificate of Calibration. The calibrated restrictor must be ordered with PN 245 059 07. Refer to: • DPFC Software Calibration with K=1 Troubleshooting Guide (PN 317 110 16) Thermo Scientific FOCUS GC - Service Manual 37 Chapter 2 Additional Technical Information Autosamplers Interface Autosamplers Interface TriPlus AS, HS, SPME, AI 3000/AS 3000 autosamplers could be installed on the FOCUS GC. Each autosampler is installed on the GC by using a dedicated bracket. Kits/Devices Option This paragraph contains technical notes about the kits and devices option available with the FOCUS GC. Oven Hydrogen Sensor The hydrogen sensor could be installed on the GC as optional device, following the installation notes reported on the dedicated operating manual. See also Hydrogen Sensor Setting in Chapter 5 of this manual. Merlin Valve Kit The Merlin Valve option (only for manual injection) could be installed on the standard S/SL injector of the GC ordering a dedicated kit PN 190 502 05. NOTE In case the GC should be equipped with a TriPlus or AI 3000 / AS3000 autosampler, in addition of the above mentioned kit, a dedicated autosampler upgrade kit must be ordered using the PN 190 502 25. More information could be found on the TriPlus or AI 3000 / AS 3000 Operating Manual. Refer to: 38 • Merlin Microseal™ Kit for S/SL and PTV Inlet System - Installation Guide (PN 317 093 65) • AI 3000/AS 3000 Operating Manual (PN 317 093 92) • TriPlus Operating Manual (PN 317 094 41) Thermo Scientific FOCUS GC - Service Manual Chapter 2 Additional Technical Information Kits/Devices Option Kit LAN for FOCUS GC This kit, PN 190 506 15, provides the materials for installing the LAN option on the GC. Refer to: • Kit LAN for FOCUS GC - Installation Guide (PN 317 094 82) Packed Column Installation On the FOCUS GC, ONLY the 1/4” and 1/8” OD metal packed columns could be installed by using the kit PN 190 500 96. Thermo Scientific FOCUS GC - Service Manual 39 Chapter 2 Additional Technical Information 40 Kits/Devices Option Thermo Scientific FOCUS GC - Service Manual Parts Replacement 3 This chapter contains detailed instructions to replace the most important mechanical and electronic parts of the FOCUS™ GC. Chapter at a Glance… Mechanical Parts Replacement.............................................................................42 Electronic Parts Replacements .............................................................................80 Operating Sequences How to Remove the Top and Rear Covers ...........................................................43 How to Replace Oven Sensor...............................................................................45 How to Replace Oven Heater ...............................................................................47 How to Replace the Oven Fan..............................................................................50 How to Replace the Oven Fan Motor...................................................................51 How to Replace the Oven Door Switch ...............................................................52 How to Replace the Air Flap Stepping Motor......................................................55 How to Replace Air Flap Spring ..........................................................................58 How to Replace the DPFC Module ......................................................................60 How to Replace the Non-DGFC Module .............................................................62 How to Replace the S/SL or PPKD Injector Assembly .......................................65 How to Replace the S/SL or PPKD Temperature Sensor and Heater ..................69 How to Replace the CB71 Detector Base Body Assembly..................................71 How to Replace the FID Temperature Sensor and Heater ...................................74 How to Replace the TCD Detector Assembly......................................................76 How to Replace the Complete Electronic Main Cassette.....................................82 How to Replace the Front Panel...........................................................................85 How to Replace MY04 and Electrometric Boards ...............................................87 Thermo Scientific FOCUS GC - Service Manual 41 Chapter 3 Parts Replacement Mechanical Parts Replacement How to Replace The CVA201 Pre-amplifier Box................................................ 88 How to Replace The TCDE203 Box.................................................................... 89 Mechanical Parts Replacement This paragraph details the instructions to replace the most important mechanical parts of the FOCUS GC. To replace the mechanical parts, please refer to the operating sequences reported in the following sections of this paragraph and to the FOCUS GC Spare Parts Catalog. NOTE • How to Remove the Top and Rear Covers on page 43. • Oven Section on page 42. • Air Flap System on page 54. • Pneumatic Modules Section on page 59. • Injector Section on page 65. • Detector Section on page 70. • Electronic Cassette Section on page 81. • Electronic Boards Section on page 86. Read the entire procedures completely to familiarize yourself with the disassembly process before proceed. Oven Section This section contains the instruction to replace the oven parts. Perform the operations referring to the relevant operating sequences. 42 • Oven Sensor, refer to How to Replace Oven Sensor on page 45. • Oven Heater, refer to How to Replace Oven Heater on page 47. • Oven Fan, refer to How to Replace the Oven Fan on page 50. • Oven Fan Motor, refer to How to Replace the Oven Fan Motor on page 51. • Oven Door Switch, refer to How to Replace the Oven Door Switch on page 52. Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Remove the Top and Rear Covers Top Cover The top cover is fixed on the main unit with two front screws and two clips. The top cover can be removed without removing the rear cover from the GC. The two front screws are visible ONLY opening the oven door (see Figure 3-1). Figure 3-1. Top Aluminium Cover Removal (1) 1. Unscrew the two screws; them remain on the main unit. 2. Manually remove top cover pulling it up from the front, the two clips will be automatically unlocked. In this mode, the rear cover can be left on the GC. Rear Cover WARNING! Before removing the covers, the GC MUST BE switched off to avoid that the over motor and the cooling fan are running during the operation. The rear cover is fixed on the main unit with four Allen screws as shown in Figure 3-2. 1. Unscrew and remove the four Allen screws that secures the rear cover by using 3-mm Allen wrench. 2. Remove the rear panel. Thermo Scientific FOCUS GC - Service Manual 43 Chapter 3 Parts Replacement Mechanical Parts Replacement 1. Carrier Gas Inlet Port 2. Detector Gas Inlet Ports 3. Servo Air Inlet port for GSV 4. Gas Sampling Valve Sample In/Out Ports 5. Oven Exhaust Vent Area 6. Electrical connections 7. Fixing Screws Figure 3-2. Rear Panel 44 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace Oven Sensor Perform the temperature sensor replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the small back cover on the rear GC bottom side unscrewing its nut. See respectively A and B of Figure 3-3. A B A. Back Cover B. Back Cover Nut Figure 3-3. Back Cover Removal (1) 4. Take note about the temperature sensor position because the new one MUST BE positioned in the same place, then loosen the Allen screw of the temperature sensor holder located in the oven (see 1 of Figure 3-4). Thermo Scientific FOCUS GC - Service Manual 45 Chapter 3 Parts Replacement Mechanical Parts Replacement 1 1. Temperature Sensor Fixing Screw Figure 3-4. Oven Temperature Sensor (1) 5. Pull the PT100 temperature sensor completely up and leave it on GC top place as shown in Figure 3-5. Figure 3-5. Oven Temperature Sensor (2) 6. From the main board disconnect the PT100 cable and remove it completely from the GC top side. 7. Install the new oven temperature sensor proceeding in the reverse order of its removal. 8. Before placing the new oven sensor in the GC oven, its fixing screw must be greased with an anti-seize lubricating compound. 46 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace Oven Heater Perform the oven heater replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the small back cover on the rear GC bottom side unscrewing its nut. See respectively A and B of Figure 3-6. A B A. Back Cover B. Back Cover Nut Figure 3-6. Back Cover Removal (2) 4. Unscrew the temperature sensor holder inside the oven. Thermo Scientific FOCUS GC - Service Manual 47 Chapter 3 Parts Replacement Mechanical Parts Replacement 5. Pull the PT100 temperature sensor completely up and leave it on GC top place. 6. Disconnect the oven heater connector and its GND cable from the main board as shown in Figure 3-7. Figure 3-7. Main Board Cables Connection (1) 7. If possible, remove the two oven heater wires from the white connector. This operation permits to pull out the oven heater cables from the small hole located on the right top corner of the oven. 8. Unscrew the four Allen screws used to fix the oven diaphragm as shown in Figure 3-8. 48 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-8. Diaphragm Screws 9. Carefully pull out the diaphragm and its heater cable. 10. Before placing the new oven heater in the GC oven, the four screws used to fix the diaphragm must be greased by using an anti-seize lubricating compound. Thermo Scientific FOCUS GC - Service Manual 49 Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the Oven Fan Perform the oven fan replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the oven heater from its oven place and leave it in the oven. 4. Unscrew the fan Allen screw and pull the fan out. In case of difficulty, an extractor should be used. 5. Install the new oven fan in the same place paying attention that 10 mm must be left between the fan and the back oven wall as shown in Figure 3-9. A home made “spacer” is suggested to have it available in the engineer’s tools kit. Figure 3-9. Correct Oven Fan Position 6. Before tightening, the screw used to fix the new oven fan on the oven motor shaft must be greased by using an anti-seize lubricating compound. 50 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the Oven Fan Motor Perform the oven motor replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the small back cover on the rear GC bottom side and the oven heater as described in How to Replace Oven Heater operating sequence on page 47. 4. Remove the oven fan as described in How to Replace the Oven Fan operating sequence on page 50. 5. Disconnect the oven motor cable from the GC main board as shown in Figure 3-10. Figure 3-10. Main Board Cables Connection (2) 6. Completely remove the four nuts used to fix the oven motor keeping it up. Thermo Scientific FOCUS GC - Service Manual 51 Chapter 3 Parts Replacement Mechanical Parts Replacement 7. Pull out the oven fan motor completely. 8. The four nuts used to fix the new oven motor, must not be tighten too much. They must be 6-7 mm equal cross screwed on their oven screws as shown in Figure 3-11. Figure 3-11. Oven Motor Nut Position 9. Reassemble all the parts proceeding in the reverse order of their removal paying attention to the position of the oven fan as described in How to Replace the Oven Fan operating sequence on page 50. 10. Before connecting the power cord to the GC, verify the oven motor function turning it manually. No mechanical noise must be heard. OPERATING SEQUENCES How to Replace the Oven Door Switch Perform the oven door switch replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the GC right panel operating as follows and referring to Figure 3-12: • 52 Unscrew the two screws A located on its internal top side. Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement • Loose the three screws B located on its external bottom side. • Disconnect the GND cable. Figure 3-12. FOCUS GC Right Panel 4. Disconnect the cable from the switch as shown in Figure 3-13. Figure 3-13. Door Switch Thermo Scientific FOCUS GC - Service Manual 53 Chapter 3 Parts Replacement Mechanical Parts Replacement 5. Unscrew the switch nut paying attention that the new switch MUST BE fixed in the same position. Air Flap System The air flap system has been developed as shown in Figure 3-14. To help the air flap system movement during its opening/closing phase, a “spring” has been used. Figure 3-14. Air Flap System The air flap system will continue to works also without the spring help. Air Flap Section This section contains the instruction to replace the air flap parts. Perform the operations referring to the relevant operating sequences. 54 • Air flap stepping motor replacement Refer to How to Replace the Air Flap Stepping Motor on page 55. • Air flap spring replacement Refer to How to Replace Air Flap Spring on page 58. Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the Air Flap Stepping Motor NOTE Air flap stepping motor is removable from the top side of the FOCUS™ GC unit. Perform the air flap stepping motor replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the GC left panel operating as follows and referring to Figure 3-15: • Unscrew the two screws A located on its internal top side. • Loose the three screws B located on its external bottom side. Figure 3-15. FOCUS GC Left Panel 4. Disconnect the stepping motor cable as shown in Figure 3-16. Thermo Scientific FOCUS GC - Service Manual 55 Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-16. Stepping Motor Cable 5. Loose the security dowel as shown in Figure 3-17. Figure 3-17. Security Dowel 6. Remove the two Allen screws used to fix the stepping motor on its place as shown in Figure 3-18. 56 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-18. Stepping Motor Screws 7. Pull up the stepping motor. CAUTION When the new stepping motor will be installed on the GC, pay attention that the shaft flat surface of the gear stepping motor, MUST BE positioned in correspondence with the security dowel as shown in Figure 3-19. It is not important if the stepping motor is not correctly positioned on its position. Stepping Motor Security Dowel Allen Screws Figure 3-19. Security Dowel Position Thermo Scientific FOCUS GC - Service Manual 57 Chapter 3 Parts Replacement Mechanical Parts Replacement 8. When the security dowel is tightened, manually rotate the stepping motor assembly up to reach its definitely position. 9. Fix the stepping motor using the two screws. 10. At this time, loose again the security dowel and then tight it again. 11. Fix the security dowel with a drop of glue. OPERATING SEQUENCES How to Replace Air Flap Spring In case of remote possibility that the air flap “spring” should be broken, perform tits replacement operating as follows: 1. Disconnect the GC rear power cord from the main power line. 2. Remove the GC top and rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the left clips pulling it out manually as shown in Figure 3-20. Figure 3-20. Air Flap Clips 4. Unscrew the Phillips screw that fix the brass bush. 58 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement 5. Using a flat-tip screwdriver, pull out the brass bush as shown in Figure 3-21. Figure 3-21. Brass Bush 6. At this time, bent out the left air flap assy and remove the spring. 7. The new spring must be one full turn (360°) pre-load before to be fixed on its position. 8. Reassemble the air flap assy proceeding in the reverse order of its removal. Pneumatic Modules Section This section contains the instruction to replace the DPFC and Non-DGFC pneumatic modules. Perform the operations referring to the relevant operating sequences. • DPFC Module replacement Refer to How to Replace the DPFC Module on page 60. • Non-DGFC module replacement Refer to How to Replace the Non-DGFC Module on page 62. Thermo Scientific FOCUS GC - Service Manual 59 Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the DPFC Module Perform the DPFC carrier module replacement operating as follows: 1. Cool down the GC at the room temperature. 2. When cooled, switch OFF the GC. 3. Close the carrier gas regulator. 4. Remove the top cover following the procedure in compliance with the GC Series as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 5. Remove the inlet tubing A from the DPFC module as shown in Figure 3-22. A. DPFC Inlet Tubing B. DPFC Module Fixing Screws Figure 3-22. DPFC Module Connections 60 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement 6. Unscrew the two Allen screws A used to fix the 3-line pre-assembled plate from the DPFC module as shown in Figure 3-23. A.3-line Pre-assembled Plate Fixing Screws B. DPFC Module Fixing Screw Figure 3-23. DPFC Module 7. Unscrew the three screws B used to fix the DGFC module on the main unit as shown in Figure 3-22 and Figure 3-23. 8. Disconnect the flat cable from the DPFC module. 9. Remove the DPFC module. 10. Install the new DPFC module proceeding in the reverse order of its removal. NOTE The DPFC module of the FOCUS GC is the same used on the TRACE GC Ultra but its holder has been mechanically updated to be fixed on the FOCUS GC. Two holes (see B of Figure 3-22) are present instead of one as the TRACE GC Ultra. In case the DPFC module for TRACE GC Ultra present in your stock should be installed on the FOCUS GC, a second hole MUST BE performed on its holder to be fixed correctly on the FOCUS GC. Thermo Scientific FOCUS GC - Service Manual 61 Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the Non-DGFC Module Perform the non-DGFC detector module replacement operating as follows: 1. Switch OFF the FID flame or the TCD filaments. 2. Cool down the GC at the room temperature. 3. When cooled, switch OFF the GC. 4. Close the detector gases regulators. 5. Remove the top cover following the procedure in compliance with the GC Series as described in the operating sequence How to Remove the Top and Rear Covers on page 43. Replace the Non-DPFC Module for FID 1. Remove the three inlet tubings A from the non-DGFC module as shown in Figure 3-24. A. Non-DGFC Inlet Tubings B. Detector Tubings Fixing Screw C. Non-DGFC Module Fixing Screws Figure 3-24. Non-DGFC Module for FID Inlet Connections 62 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement 2. From the non-DGFC module, disconnect the three cables identified as Air, Hydrogen and Make-up as shown in Figure 3-25. A. Air Cable B. Hydrogen Cable C. Make-up Cable Figure 3-25. Non-DGFC Module for FID Cables Connections 3. From the non-DGFC module, unscrew the Allen screw B used to fix the 3-line pre-assembled plate as shown in Figure 3-24. 4. Remove the two Allen screws C used to fix the non-DPFC module on the GC main unit as shown in Figure 3-24. 5. Install the new non-DGFC module proceeding in the reversed order of its removal. Replace the Non-DPFC Module for TCD 1. Remove the inlet tubing A from the non-DGFC module as shown in Figure 3-26 on page 64. Thermo Scientific FOCUS GC - Service Manual 63 Chapter 3 Parts Replacement Mechanical Parts Replacement C B A A. Non-DGFC Inlet Tubing B. Detector Tubings Fixing Screw C. Non-DGFC Module Fixing Screw Figure 3-26. Non-DGFC Module for TCD Inlet Connections 2. From the non-DGFC module, disconnect the two cables identified as Reference and Make-up as shown in Figure 3-27. R M R. Reference Cable M. Make-up Cable Figure 3-27. Non-DGFC Module for TCD Cables Connections 64 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement 3. From the non-DGFC module, unscrew the Allen screw B used to fix the 3-line pre-assembled plate as shown in Figure 3-26. 4. Remove the Allen screw C used to fix the non-DPFC module on the GC main unit as shown in Figure 3-26. 5. Install the new non-DGFC module proceeding in the reverse order of its removal. Injector Section This section contains the instruction to replace the injector parts. Perform the operations referring to the relevant operating sequences. • S/SL injector assembling replacement Refer to How to Replace the S/SL or PPKD Injector Assembly on page 65. • Injector temperature sensor and heater replacement Refer to How to Replace the S/SL or PPKD Temperature Sensor and Heater on page 69. OPERATING SEQUENCES How to Replace the S/SL or PPKD Injector Assembly Perform the injector assembly replacement operating as follows: 1. Cool down the GC at the room temperature. When cooled, switch OFF the GC. 2. If installed, remove the column from the injector. 3. Remove the top cover following the procedure in compliance with the GC Series as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 4. Remove the two Phillips screws used to fix the insulating material cover shown in Figure 3-28. Thermo Scientific FOCUS GC - Service Manual 65 Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-28. Insulating Material Cover 5. Unscrew the two Allen screws used to fix the 3-line pre-assembled plate on the DPFC module. See A of Figure 3-23 in the How to Replace the DPFC Module operating sequence on page 60. 6. Remove the insulating material around the injector. 7. Remove the injector septum cap. 8. Remove the screw 1 and the aluminium ring 2 that stops the temperature sensor and the heater element in the heating block as schematically shown in Figure 3-29 on page 67. 66 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement 1. Screw 2. Aluminium Ring Figure 3-29. Injector Temperature Sensor and Heater Screw 9. Take out the temperature sensor and the heater from the heating block and leave them on the top of the GC. 10. Remove the small nut and the small metal cover with the insulator material inside the oven as shown in Figure 3-30. Figure 3-30. Nut Inside the Oven Thermo Scientific FOCUS GC - Service Manual 67 Chapter 3 Parts Replacement Mechanical Parts Replacement 11. Unscrew the nut of the column adapter (see 1 and 2 of Figure 3-31) and remove it from the injector. 1. Column Adapter Nut 2. Column Adapter 3. Blocking Nut Figure 3-31. Column Adapter 12. Unscrew the injector blocking nut (see 3 of Figure 3-31) and pull the injector out from its housing. 13. Install the new injector proceeding in the reverse order of its removal paying attention that the “silver seal”, present on the column adapter, must be changed with a new one having PN 290 026 29. 14. Before placing the temperature sensor and the heater into the injector heating block, grease them by using an anti-seize lubricating compound. 68 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the S/SL or PPKD Temperature Sensor and Heater Perform the replacement of the injector temperature sensor and heater operating as follows: 1. Perform the steps from 1 to 9 of the How to Replace the S/SL or PPKD Injector Assembly operating sequence, on page 65, up to reach the removing phase of the temperature sensor and heater from heating block. 2. Remove the small back cover on the rear GC bottom side unscrewing its nut. See respectively A and B of Figure 3-32. A B 1. Back Cover 2. Back Cover Nut Figure 3-32. Back Cover Removal (3) Thermo Scientific FOCUS GC - Service Manual 69 Chapter 3 Parts Replacement Mechanical Parts Replacement 3. Disconnect the relative cable (P6 injector heater - P10 injector temperature sensor) from the MY04 main board, then remove the interested item from the GC as shown in Figure 3-33. Figure 3-33. Main Board Cables Connection (3) 4. Before placing the temperature sensor into the injector heating block, grease it by using an anti-seize lubricating compound. 5. Reassemble all the other parts proceeding in the reverse order of their removal. Detector Section This section contains the instruction to replace the detector parts. Perform the operations referring to the relevant operating sequences. 70 • FID Detector assembling replacement Refer to How to Replace the CB71 Detector Base Body Assembly on page 71. • FID temperature sensor and heater replacement Refer to How to Replace the FID Temperature Sensor and Heater on page 74. • TCD assembling replacement Refer to How to Replace the TCD Detector Assembly on page 76. Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the CB71 Detector Base Body Assembly Perform the CB71 detector assembly replacement operating as follows: 1. Cool down the GC at the room temperature. 2. When cooled, switch OFF the GC. 3. Close the detector gas regulators. 4. Remove the GC top cover following the procedure in compliance with the GC Series as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 5. Remove the column from the base body. 6. Remove the FID detector from the base body. 7. Remove the two Phillips screws used to fix the CB71 detector insulating material cover as shown in Figure 3-34. Figure 3-34. Detector Insulating Material Cover 8. Unscrew the Allen screw A used to fix the 2-line pre-assembled plate on the non-DGFC module as shown in Figure 3-35. Thermo Scientific FOCUS GC - Service Manual 71 Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-35. 3-Line Pre-assembled Plate Fixing Screw 9. Remove the capillary column adapter from the capillary detector base body. 10. Remove the detector base body fixing nut inside the oven as shown in Figure 3-36. Figure 3-36. Detector Base Body Fixing Nut 11. Remove the base body assembly pulling it up paying attention at the temperature sensor and heater cables. 12. Loose the relevant fixing screw to remove the temperature sensor and the heater from the detector heating block as shown in Figure 3-37 on page 73. 72 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-37. Heater-Temperature Sensor Fixing Nut 13. Take out the temperature sensor and the heater from the detector heating block. 14. Before placing the temperature sensor and the heater into the detector heating block, grease them by using an anti-seize lubricating compound. 15. Install the new CB71 detector base body proceeding in the reverse order of its removal. Thermo Scientific FOCUS GC - Service Manual 73 Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the FID Temperature Sensor and Heater Perform the replacement of the FID temperature sensor and heater operating as follows: 1. Perform the steps from 1 to 7 of the How to Replace the CB71 Detector Base Body Assembly operating sequence, on page 71, up to reach the temperature sensor and heater removing phase from detector base body heating block. 2. Remove the small back cover on the rear GC bottom side unscrewing its nut. See respectively A and B of Figure 3-38. A B 1. Back Cover 2. Back Cover Nut Figure 3-38. Back Cover Removal (4) 3. Disconnect the relative cable (P7 detector heater - P9 detector temperature sensor) from the MY04 main board, then remove the interested item from the GC as shown in Figure 3-39. 74 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-39. Main Board Cables Connection (3) 4. Before placing the temperature sensor and the heater into the detector heating block, grease them by using an anti-seize lubricating compound. 5. Reassemble all the other parts proceeding in the reverse order of their removal. Thermo Scientific FOCUS GC - Service Manual 75 Chapter 3 Parts Replacement Mechanical Parts Replacement OPERATING SEQUENCES How to Replace the TCD Detector Assembly The TCD detector assembly consists of a properly insulating box containing the TCD cell as shown in Figure 3-40. WARNING! The replacement of the TCD cell, temperature sensor and heaters, MUST be performed only by Thermo Fisher Scientific EXPERT field service engineers. TCD Detector Assembly with Cover TCD Detector Assembly without Cover Figure 3-40. TCD Detector Assembly Perform the complete TCD detector assembly replacement operating as follows: 1. Cool down the GC at the room temperature.and turn OFF the filaments of the TCD. 2. When cooled, switch OFF the GC. 3. Close the detector gas regulators. 4. Remove the detector and the GC top covers following the procedure in compliance with the GC Series as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 76 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Mechanical Parts Replacement 5. Remove the column and the column adapter, if present, from the detector base body. 6. Unscrew the Allen screw A used to disconnect the 2-line pre-assembled plate from the non-DGFC module as shown in Figure 3-41. A Figure 3-41. “-Line Pre-assembled Plate Removal) 7. Remove the small back cover on the rear GC bottom side unscrewing its nut. See respectively A and B of Figure 3-42. A B 1. Back Cover 2. Back Cover Nut Figure 3-42. Back Cover Removal (5) 8. On the MY04 main board disconnect the detector heater and the temperature sensor cables, respectively from connectors marked P7 and P9, then remove the interested item from the GC as shown in Figure 3-43. Thermo Scientific FOCUS GC - Service Manual 77 Chapter 3 Parts Replacement Mechanical Parts Replacement Figure 3-43. Main Board Cables Connection (4) 9. On the top of the TCDE 203 box, located on the GC rear, disconnect the heater and filaments cables respectively from the connectors marked P7 and P3. 10. Unscrew the four screws that fix the TCD detector assembly to the injector/ detector compartment as shown in Figure 3-44 1 3 1 2 1. TCD Detector Assembly 2. TCD Detector Assembly Fixing Screws 3. Ground Connection Figure 3-44. TCD Detector Assembly Removal 78 Thermo Scientific FOCUS GC - Service Manual Mechanical Parts Replacement Chapter 3 Parts Replacement 11. Disconnect the ground wire, coming from the TCD detector assembly from the ground connection located on the upper part of the GC behind the injector/ detector compartment. See Figure 3-44. 12. Remove the TCD detector assembly. 13. Install the new TCD detector assembly proceeding in the reverse order of its removal. Thermo Scientific FOCUS GC - Service Manual 79 Chapter 3 Parts Replacement Electronic Parts Replacements Electronic Parts Replacements This paragraph details the instructions to replace the most important electronic parts of the FOCUS GC. Introduction According to GC version refer to: • GC with FID • GC with TCD GC with FID In the GC there are following electronic boards: 1. The main board identified as MY04 is used to control the GC functions. 2. A second small board identified as ELI201 is used to control the FID detector electrometric functions. These two boards are located in the lower part of the GC into a dedicated cassette. The boards must be removed together. The cassette is removable from the front of the GC unscrewing the screw located on the rear panel, disconnecting all the cables connected on the main board and removing the GC LEFT cover panel to have the FID cables free. In addition to the ELI201 board, another circuit, identified as CVA201 (Electrometer pre-amplifier box) is used to read the signal coming from the FID detector. The ELI201 board is fixed on the main board while the CVA201 electrometer preamplifier circuit is positioned on the rear of the GC into a box. To replace the electronic parts, please refer to the operating sequences reported in the following sections of this paragraph. • 80 Electronic Cassette Section on page 81. Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Electronic Parts Replacements • NOTE Electronic Boards Section on page 86. Read the entire procedures completely to familiarize yourself with the disassembly process before proceeding. GC with TCD In the GC there are the following electronic boards: 1. The main board identified as MY04 is used to control the GC functions. 2. A second board identified as TCDI 203 is used to control the TCD detector electrometric functions. These two boards are located in the lower part of the GC into a dedicated cassette. The boards must be removed together. The cassette is removable from the front of the GC unscrewing the screw located on the rear panel, disconnecting all the cables connected on the main board and removing the GC LEFT cover panel to have the detector cables free. In addition to the TCDI 203 board, another circuit, identified as TCDE 203 is used to read the signal coming from the TCD detector. The TCDI 203 board is fixed on the main board while the TCDE 203 board is positioned on the rear of the GC into a box. To replace the electronic parts, please refer to the operating sequences reported in the following sections of this paragraph. NOTE • Electronic Cassette Section on page 81. • Electronic Boards Section on page 86. Read the entire procedures completely to familiarize yourself with the disassembly process before proceeding. Electronic Cassette Section This section contains the instruction the replace the electronic cassette parts. Perform the operations referring to the relevant operating sequences. Thermo Scientific FOCUS GC - Service Manual 81 Chapter 3 Parts Replacement Electronic Parts Replacements • Complete electronic cassette replacement Refer to How to Replace the Complete Electronic Main Cassette on page 82. • Front panel replacement Refer to How to Replace the Front Panel on page 85. OPERATING SEQUENCES How to Replace the Complete Electronic Main Cassette Perform the complete electronic main cassette replacement operating as follows: 1. Cool down the GC at the room temperature. 2. When cooled, switch OFF the GC. 3. Disconnect the GC power cord. 4. Remove the top cover following the procedure in compliance with the GC Series as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 5. Remove the GC left panel operating as follows and referring to Figure 3-45 on page 83. 82 • Unscrew the two screws A located on its internal top side • Loose the three screws B located on the external bottom side of it and remove its GND cable. Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Electronic Parts Replacements Figure 3-45. FOCUS™ GC Left Panel 6. Remove the small back cover on the GC rear bottom side unscrewing its nut. See respectively A and B of Figure 3-46. A. Back Cover B. Back Cover Nut C. Electronic Main Cassette Fixing Screw Figure 3-46. Back Covers and Electronic Main Cassette Screw 7. Disconnect all the cables from the main board. 8. Unscrew the Allen screw C fixed on the GC rear side as shown in Figure 3-46. Thermo Scientific FOCUS GC - Service Manual 83 Chapter 3 Parts Replacement Electronic Parts Replacements 9. In case of FID, remove the excitation/polarization cable from the main unit, as shown in Figure 3-47, and locate it into the rear electronic main cassette. From the CVA201 pre-amplifier box disconnect the small flat cable and locate it into the electronic main cassette too. Figure 3-47. FID Cables 10. In case of TCD, from the TCDE 203 box disconnect the flat cable and locate it into the electronic main cassette too. WARNING! The cables must be located into the electronic main cassette in a way to permit to it to be easily removed without any obstacle. 11. Carefully pull out completely the electronic main cassette from the front of the GC. WARNING! Do not insist to pulling out electronic main cassette if some mechanical obstacles are encountered. 84 Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Electronic Parts Replacements OPERATING SEQUENCES How to Replace the Front Panel The GC display and its keypad are not available as single spare parts. In case of troubles on them, the complete Front Panel must be replaced. Perform the front panel replacement operating as follows: 1. Cool down the GC at the room temperature. 2. When cooled, switch OFF the GC. 3. Disconnect the GC rear power cord from the main power line. 4. Disconnect all the other cables connected to the GC rear panel. 5. Remove the small back cover on the GC rear bottom side unscrewing its nut. See respectively A and B of Figure 3-48. 1. Back Cover 2. Back Cover Nut C. Electronic Main Cassette Fixing Screw Figure 3-48. Back Covers and Electronic Cassette Screw (2) 6. Unscrew the Allen screw C fixed on the GC rear panel as shown in Figure 3-48. The electronic main cassette will be pulled carefully out for a few centimeters. Thermo Scientific FOCUS GC - Service Manual 85 Chapter 3 Parts Replacement Electronic Parts Replacements 7. By using a short Phillips screwdriver, remove the five screws A used to fix the Front Panel of the electronic main cassette as shown in Figure 3-49. A. Front Panel Fixing Screws B. Front Panel Cable Figure 3-49. Front Panel Internal View 8. Remove the front panel flat cable B from the mother board as shown in Figure 3-49. Electronic Boards Section This section contains the instructions the replace the electronic boards. Three standard electronic equipments are present on the GC. They are identified as follows: FOCUS GC with FID • MY04 pcb (Main board) • ELI201 pcb (FID electrometric board) • CVA201pcb (Pre-amplifier box) FOCUS GC with TCD 86 • MY04 pcb (Main board) • TCDI203 pcb electrometric board) • TCDE203 pcb box Thermo Scientific FOCUS GC - Service Manual Chapter 3 Parts Replacement Electronic Parts Replacements The main MY04 board and the ELI201 or TCDI203 boards CAN NOT BE removed individually. Both are located in the electronic main cassette. Perform the operations referring to the relevant operating sequences. • MY04 and electrometric boards replacement Refer to How to Replace MY04 and Electrometric Boards on page 87. • CVA201 replacement Refer to How to Replace The CVA201 Pre-amplifier Box on page 88. • TCDE203 box replacement Refer to How to Replace The TCDE203 Box on page 89 OPERATING SEQUENCES How to Replace MY04 and Electrometric Boards Perform the MY04 and ELI201 or TCDI203 boards replacement operating as follows: Before starting, remove completely the electronic main cassette as described in How to Replace the Complete Electronic Main Cassette operating sequence. Electrometric Board Replacement 1. No screws are used to fix the ELI201 or TCDI203 on the MY04 main board. 2. The electrometer board is connected to the main board through the connector marked P30. Remove the board pulling it out manually from its connector. MY04 Main Board Replacement Referring to Figure 3-50, operate as follows. 1. Disconnect the Front Panel cable from the connector P28. 2. Remove the Time Events cable from the connector P18. 3. Remove the four RS232 connector’s screws from the MY04 board holder. Thermo Scientific FOCUS GC - Service Manual 87 Chapter 3 Parts Replacement Electronic Parts Replacements 4. Remove the twelve screws, marked as red bullets in 3-50, used to fix the main board on its bracket. Figure 3-50. MY04 Main Board OPERATING SEQUENCES How to Replace The CVA201 Pre-amplifier Box Perform the CVA201 Pre-amplifier box replacement operating as follows: 1. Set OFF the FID flame and cool down the GC. 2. When cooled, switch OFF the GC and remove its power cord. 3. Remove the GC rear cover as described in Chapter 2 of this manual. 4. Remove also the GC left cover panel removing the two screws located on the top of the GC and unscrewing the three screws located on the bottom side of the left panel (see Figure 3-15 on page 55). 5. From the FID detector cell disconnect the detector signal cable coming from the CVA201 box. 88 Thermo Scientific FOCUS GC - Service Manual Electronic Parts Replacements Chapter 3 Parts Replacement 6. From the CVA201 box disconnect the small flat cable coming from the ELI201 board. 7. Remove the CVA201 box from its holder clips. OPERATING SEQUENCES How to Replace The TCDE203 Box Perform the TCDE203 box replacement operating as follows: 1. Set OFF the TCD filaments and cool down the GC. 2. When cooled, switch OFF the GC and remove its power cord. 3. Remove the GC rear cover as described in Chapter 2 of this manual. 4. Remove also the GC left cover panel removing the two screws located on the top of the GC and unscrewing the three screws located on the bottom side of the left panel (see Figure 3-15 on page 55). 5. On the top of the TCDE203 box, disconnect the heater and filaments cables coming from the TCD detector assembly. 6. On the bottom of the TCDE203 box disconnect the flat cable coming from the TCDI203 board, the 48 Vac and the GND cables. 7. Remove the TCDE203 box undoing the two screws that fix it to the upper part of the GC as shown in Figure 3-51 on page 90. Extract the box from the back of the GC pay attention to the duct for screwdriver. Thermo Scientific FOCUS GC - Service Manual 89 Chapter 3 Parts Replacement Electronic Parts Replacements 2 1 Box Fixing Screws 1 2. Duct for Screwdriver Figure 3-51. TCDE203 Box 90 Thermo Scientific FOCUS GC - Service Manual Electronic Boards Description 4 This chapter contains a detailed description of the electronic boards of the FOCUS™ GC. Chapter at a Glance… General Description..............................................................................................92 Electronic Main Cassette......................................................................................94 Main Schematic Diagrams ...................................................................................96 Introduction to MY04 Main Board.......................................................................97 Connectors Dismounting on Main MY04 Board ...............................................106 MY04 Main Board Sections Description ...........................................................108 ELI201 Electrometric Board (FID) .................................................................... 113 CVA201 Electrometer Pre-amplifier Board (FID) ............................................. 115 TCDI203 Electrometric Board (TCD)................................................................ 116 TCDE203 Board (TCD) ..................................................................................... 117 FRTMY100 Board and Keypad.......................................................................... 118 IFM311 LAN Option.......................................................................................... 119 Operating Sequences How to Dismounting the Connectors .................................................................106 Thermo Scientific FOCUS GC - Service Manual 91 Chapter 4 Electronic Boards Description General Description General Description The FOCUS™ GC is subdivided in five main compartments: • Oven • Injector/Detector • Digital Pressure Flow Control (DPFC) for carrier gas • Manual Flow Control (Non-DGFC) for detector gases • Electronic Main Cassette The following parts are located in the Oven compartment: • Oven fan • Oven motor • Oven heater • Oven temperature sensor The following parts are located in the injector/detector compartment according to the GC configuration: • One split-splitless (S/SL) injector or purged packed column injector (PPKD) • One base-body for FID detector or TCD detector assembly • Heater cartridges • PT100Temperature sensors The following parts are located in the DPFC and non-DGFC compartment present on the upper parts of the GC rear side: • One DPFC pneumatic module for the digital carrier gas control • One non-DGFC conventional pneumatic module for the detector gases control The following parts are located in the electronic main cassette • 92 The main board which contains the following parts: Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description General Description — Main input voltage components (breaker, filter, fuses and others — Rear connectors — earth connections — power supply — SSR devices — digital/analogical — microprocessors NOTE Refer to the FOCUS GC Spare Parts Catalog for the available spare parts Thermo Scientific FOCUS GC - Service Manual 93 Chapter 4 Electronic Boards Description Electronic Main Cassette Electronic Main Cassette The primary requirement for any FOCUS™ GC is to have multiple independent temperature controlled zones. In addition, various input and output signals are needed for synchronization and control of peripheral devices as heaters, temperature sensors, solenoid valves, AC motor, stepping motor and electrical switch. In the GC, all functions are controlled by two microprocessors present in the Main Board of the Main Cassette located under the GC Oven. Figures 4-1 and 4-2 show a generalized block diagrams of the GC with FID and TCD respectively. RS 232 (PC) MAIN C.P.U. RS 232 (AS) Timed Events Handshake/Sampler Power Supply Valve Drivers +5 Vdc ; +/- 15 Vdc 24 Vdc Valves ELI201 Electrometric Board DPFC Module DPFC + A/D + D/A. H2 Sensor Oven Flap Stepping Motor Heater Drivers Temp. Control Heaters (230/120 Vac) Oven Door Switch Temperature Sensor CVA 210 Electrometer Pre-amplifier FID Detector Analogical Output Voltages Figure 4-1. FOCUS™ GC with FID Main Board Block Diagram 94 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Electronic Main Cassette RS 232 (PC) MAIN C.P.U. RS 232 (AS) Timed Events Handshake/Sampler Power Supply Valve Drivers +5 Vdc ; +/- 15 Vdc 24 Vdc Valves TCDI203 Electrometer Board DPFC + A/D + D/A. Heater Drivers Heaters (230/120 Vac) Oven Door Switch TCDE203 Box DPFC Module H2 Sensor Oven Flap Stepping Motor Temp. Control Temperature Sensor TCD Detector Analogical Output Voltages Figure 4-2. FOCUS™ GC with TCD Main Board Block Diagram The Main Board is divided in six main functions named: • Power Supply • Main Central Processor Unit CPU • Analog/Digital & DPFC • Temperature Controller • Heater Drivers • Valve drivers Thermo Scientific FOCUS GC - Service Manual 95 Chapter 4 Electronic Boards Description Main Schematic Diagrams Main Schematic Diagrams The schematics and component layout information is listed below in Table. The number of drawings has been included for each schematic and its corresponding layout. The wiring diagram and the block diagram shown you all internal cables and connectors used to connect all heaters, temperature sensors, boards, motors and transformers. Table 4-1. Main Schematic Diagrams FOCUS GC Configuration 96 Schematic Diagram Layout Wiring Diagram With FID detector SE 1529 SC 817 SC 818 With TCD detector SE 1528 SC 819 SC 820 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Introduction to MY04 Main Board This board provides the platform for communication between many of the other PCB’s functions of the main board of the FOCUS™ GC and the external devices. 1 1. MY04 Main Board 2. Transformer (+ 5V; 700 mA) 3. Transformer (±15 V; 900 mA) 4. Transformer (+24 V; 500 mA) 5. Eprom 6. Reset Button 7. RS232 Connectors 8. Detector Board Connector (NOT USED) 9. Microprocessor 10. Time Events Connectors 11. DPFC CPU Section 12. Main CPU Section 13. DPFC Connector 14. Back-up Battery Figure 4-3. MY04 Main Board The following tables list the protection fuses and the connectors present on the MY04 Main Board. Thermo Scientific FOCUS GC - Service Manual 97 Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Table 4-2. Fuses Fuses F1 - F3 F2 F4 - F5 Description Injector/Detector heaters Transformers (on primary coils) Oven motor Table 4-3. Connectors Connectors Description P2 120V Voltage Selector P3 230V Voltage Selector P4 Fan Motor P6 Injector Heater P7 Detector Heater The above individual Receptacle Pod connectors are collected in a single Receptacle Shroud connector to be easily removed. In case an individual Receptacle Pod connector should be removed, please refer to the Connectors dismounting on Main MY04 Electronic board. P5 Oven Heater P8 Oven Temperature Sensor P9 Detector Temperature Sensor P10 Injector Temperature Sensor P19 Door Switch P20 Flap Stepping Motor The above individual Receptacle Pod connectors are collected in a single Receptacle Shroud connector to be easily removed. In case an individual Receptacle Pod connector should be removed, please refer to the Connectors dismounting on Main MY04 Electronic board. 98 P13 - P14 Split Purge Valves P15 ÷ P17 Manual Detector Module P24 DPFC Connection P27 Free Connector for future expansion P30 FID/TCD (ELI201/TCDI203) Board Connector Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Power Supply Notes This board has been developed to operate without any electronic modification with 120V or 230V main power line. Only the P2 and P3 connectors must be positioned/connected on the main board in accordance with the main power line value. The Injector, Detector and Oven heaters, the Oven Fan motor (with its capacitor) and the Main Power Switch are factory set in accordance with the FOCUS™ GC voltage required. General Notes The main board is divided in CPU, power and buffer sections. The CPU section is divided in two subsections where the main controls all the GC functions and the second controls the DPFC module function. The two subsections are always in communication between them. The temperature control for injector, detector base body and GC oven use the temperature sensors (PT100) as feedback elements. All of these signals are routed through the temperature controller section. The main board has: • A 32+32 pins “Piggy Back” connector to accept the electrometer board. • Five connectors positioned to provide control signals to external peripherals via the rear panel of the GC. These connectors are: — Time Event — Generic Handshake — Sampler Signal — RS 232 - Autosampler — RS 232 - Computer Timed Event Connector The following Figure 4-4 shows the Timed Events connector (P18, 15 D-pin Female) which provides generic timed control signals. Thermo Scientific FOCUS GC - Service Manual 99 Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Figure 4-4. Front View of Timed Event Connector Individual pin assignments for the Timed Event connector, and associated control signals and electrical characteristics, are indicated in Table 4-4. Table 4-4. Timed Event Connector 100 Pin Number Signal Electrical Characteristics 1 Aux 1-vlv Output Driver (direct) 24 V, 1 W Max 2 Aux 1-vlv Output Driver (direct) 24 V, 1 W Max 3 Aux 2-vlv Output Driver (direct) 24 V, 1 W Max 4 Aux 2-vlv Output Driver (direct) 24 V, 1 W Max 5 Sense H2 Input Pull-up 5 Vdc (for Open Collector Output) 9 PH0 Open Collector, 24 V Max, 15 mA Current Sink 10 PH2 Open Collector, 24 V Max, 15 mA Current Sink 11 PH4 Open Collector, 24 V Max, 15 mA Current Sink 12 PH8 Open Collector, 24 V Max, 15 mA Current Sink 6,7,8,13,14,15 GND Ground • Two power Auxiliary signals are available to control two solenoid valves directly. • The “Sense H2” is an inlet signal coming from the Hydrogen Sensor Device when installed. Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Introduction to MY04 Main Board • The four Open Collector signals allows to recognize the GC status phases. The signals are present in hexadecimal code. The Code/Phase are reported in the Table 4-5. — In the GC there are seven selectable RAMPS (°C/min) and eight selectable ISOTHERMS (temp./time). Their will be named as INITIAL, MIDDLE and FINAL.: Table 4-5. Code-Phases Code Phase Code Phase 0 Not Ready 8 Middle Rate 1 ST/BY 9 Not Used 2 Ready to Inject 10 Middle Iso 3 Not Used 11 Final Rate 4 Initial Iso 12 Not Used 5 Initial Rate 13 Final Iso 0 Not Ready 8 Middle Rate 1 ST/BY 9 Not Used Generic Handshake Connector Figure 4-5 shows the Generic Handshake connector (P21, MINIDIN 8-Pin Female Circular), which provides generic handshake input/output control signals. . Figure 4-5. Front View of Generic Handshake Connector Individual pin assignments for the Generic Handshake connector, and associated control signals and electrical characteristics, are indicated Table 4-6. Thermo Scientific FOCUS GC - Service Manual 101 Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Table 4-6. Generic Handshake Connector Pin Number Electrical Characteristics Signal 1 Remote Start—This signal is an input to the GC to start the program temperature. Polarity is software selectable. TTL Input 2 Inhibit Ready—This signal is an input to the GC to indicate that the GC is not ready to start a run. Polarity is software selectable. TTL Input 3 Ground 4 Start Run—This signal is an output from the GC to indicate external devices that a run has started. Polarity is software selectable. Open Collector, 24 V Max, 15 mA Output Current Sink 5 Ready Out—This signal is an output from the GC to indicate external devices that the GC is ready to start a run. Polarity is software selectable. Open Collector, 24 V Max, 15 mA Output Current Sink 6 End Run—This signal is an output from the GC to indicate external devices that the GC run has ended. Polarity is software selectable. Open Collector, 24 V Max, 15 mA Output Current Sink 7 Prep Run—This signal is an output from the GC to signal external devices that the GC is in the Prep Run condition. Polarity is software selectable. Open Collector, 24 V Max, 15 mA Output Current Sink 8 Not Used N/A --- Sampler Signal Connector Figure 4-6 shows the Sampler Signal connector (P22, MINIDIN 6-Pin Female Circular), which provides autosampler input/output control signals 102 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Introduction to MY04 Main Board . Figure 4-6. Front View of Sampler Signal Connector Individual pin assignments for the Sampler Signal connector, and associated control signals and electrical drive capabilities, are indicated in Table 4-7. Table 4-7. Autosampler Connector Pin Number Signal Electrical Drive Capability 1 Remote Start—This signal is an input to the GC to indicate that the run should start. Polarity is software selectable. TTL Input 2 Inhibit Ready—This signal is an input to the GC to indicate that the GC should not go Ready. Polarity is software selectable. TTL Input 3 Ground 4 Start Run—This signal is an output from the GC to indicate external devices that a run has started. Polarity is software selectable. Open Collector, 20 V Max, 150 mA Current Sink 5 Ready Out—This signal is an output from the GC to signal external devices that the GC is Ready to start a run. Polarity is software selectable. Open Collector, 20 V Max, 150 mA Current Sink 6 Not Used N/A Thermo Scientific FOCUS GC - Service Manual --- 103 Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Autosampler Connector Figure 4-7 shows the Autosampler connector (P23, MINIDIN 9-Pin Female Standard), which provides the RS 232 serial line communication signals to/from the autosampler. 1 5 9 6 Figure 4-7. Autosampler Connector Individual pin assignments for the Autosampler connector, and associated control signals and electrical characteristics, are indicated Table 4-8. Table 4-8. Autosampler Connector Pin Number Signal Electrical Characteristics 2 TXD Transmitted Data 3 RXD Received Data 4 DTR Data Terminal Ready 5 GND Ground 8 CTS Clear To Send Computer Connector Figure 4-8 shows the Computer connector (P29, MINIDIN 9-Pin Female Standard), which provides the RS 232 serial line communication signals to/from the computer. 1 5 9 6 Figure 4-8. Computer Connector 104 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Introduction to MY04 Main Board Individual pin assignments for the Autosampler connector, and associated control signals and electrical characteristics, are indicated Table 4-9. Table 4-9. Computer Connector Pin Number Signal Electrical Characteristics 2 TXD Transmitted Data 3 RXD Received Data 4 DTR Data Terminal Ready 5 GND Ground 8 CTS Clear To Send Other connectors provide inter chassis power distribution and communication. These connectors are itemized in Table 4-10. Table 4-10. Other Connectors Connector P28 Thermo Scientific FOCUS GC - Service Manual Function Communication with Keypad and Display Panel 105 Chapter 4 Electronic Boards Description Connectors Dismounting on Main MY04 Board Connectors Dismounting on Main MY04 Board To simplify the connections/disconnections on the MY04 board of some connectors, they have been placed in a single Receptacle connector. OPERATING SEQUENCES How to Dismounting the Connectors In case it is necessary to take out the individual small Receptacle Pod connector from the single main Receptacle connector, operate as follows: 1. Disconnect the FOCUS GC rear power cord from the main power line. A B 1. Back Cover 2. Back Cover Nut Figure 4-9. Back Cover 2. Remove the GC top cover and/or the rear covers as described in the operating sequence How to Remove the Top and Rear Covers on page 43. 3. Remove the small back cover on the rear GC bottom side (see A of Figure 4-9), unscrewing its blocking nut (see B of Figure 4-9). 106 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description Connectors Dismounting on Main MY04 Board 4. Take out the interested receptacle connector pressing the two lateral sides and then pulling out it as shown in Figure 4-10. A. Lateral Side B. Bottom View Figure 4-10. Receptacle Shroud 5. To remove a receptacle pod, shown in Figure 4-11, from its receptacle connector, use a small screwdriver to lift the respective receptacle pod notch, then pull out it. Receptacle Pod Pod Notch Figure 4-11. Receptacle Pod Thermo Scientific FOCUS GC - Service Manual 107 Chapter 4 Electronic Boards Description MY04 Main Board Sections Description MY04 Main Board Sections Description This paragraph provides the description of the main section of the MY04 main board. Schematics and Component Layout The schematics and component layout information for the MY04 board is listed below in Table 4-11. The number of drawings has been included for each schematic and its corresponding layout. Table 4-11. MY04 Main Board Schematics Board MY04 PN Drawing Layout 236 485 17 SE 1526 (1/7) PC 1206 Sheet 1/7 (CPU 1) The CPU PCB provides the control for all functions of the FOCUS™ GC. The main 8 bits microprocessor is ZILOG Z180 (U52) having internally: • One megabyte of address memory • Two RS232 • One SPI (Serial Peripheral Interface) • Three Timers There are two types of memories selected by 74HC00 (U46): • Flash memory (embedded firmware), 128 kilobyte (U51) • Static memory, 128 kilobyte (U53) The oscillation frequency is of 12,288 MHz generated from Quartz (Y2) and then it is divided for two by Microprocessor. The integrated circuit MAX705 (U71) generates the RESET logic following the +5Vdc. 108 Thermo Scientific FOCUS GC - Service Manual MY04 Main Board Sections Description Chapter 4 Electronic Boards Description The FRAM 25C160-S (U69) is used for “store configuration” loading memory only. A real time clock circuit (used for time and date) and interrupt are generated from DS1305E (U73) which is supplied also by a 3.6 V backup battery (BAT1). The oscillation frequency is of 32.768 kHz generated from Quartz Y3. Two Is MAX232CPE (U58-U59) drive two serial communication outputs (PC-AS) and the LAN module interface IFM 311. The chips selected are buffered by ICs 74HC377 (U54-U57) and 74HC245 (U65). Sheet 2 (CPU2) The FID connector marked P30 is used for the electrometer board ELI201 or TCDI203, incorporated in the GC, which are respectively connected to: • the pre-amplifier module CVA201 and then to the FID detector cell. • the TCDE203 box and then to the TCD detector cell Two ICs 74HC590 (U61-U60) read the 0-3.5 MHz input frequency (baseline signal) generated from the detector card (FREQ). The TERMINAL BLOCK (ANALOG OUT) provide the analogical output voltages (0-1V; 0-10V) for an external connection to Integrator or Recorder. Two ICs 74HC160 (U62) and 74HC4020 (U67) are the counter circuit with 1/300 sec. (f = 300 Hz) of scansion. A dedicated interrupt circuit (INT0) is designed by ICs 74HC74 (U45). The display and keyboard panel is connected to main board through the P28 connector (DISPLAY CONN.). The output signals present on rear panel connector (e.g.: Generic Handshake, Time Events, etc…) are generated by two ICs LM339 (U55-U56) as “open collector” signals. The integrated circuits 74HC4020 (U67), 74HC126 (U63) and BC337 (Q11) control the Buzzer device (BZ1). Thermo Scientific FOCUS GC - Service Manual 109 Chapter 4 Electronic Boards Description MY04 Main Board Sections Description Sheet 3 (DPFC1) The single-chip 8 bits microprocessor MOTOROLA 68HC11 (U25), supplied from 8 MHz oscillator (Y1), controls: • The Analog/Digital circuit (U28, DG508A) used for Digital Pressure Flow Control (DPFC for carrier gas) • The Analogic/Digital circuit (U23, DG508A) used for temperature controls (Oven, Detector Base Body and Injector) Two ICs 74HC244 (U16-U17) read all inputs (high temp/low limits, H2 sensor, etc.) which could be generated the “safety cut off “on the GC (alarm status). The circuit with 74HC00 (U30) and 74HC139 (U36) provides three chip-selects (WR1, BOOT, RAM) to able the 128 Kilobytes RAM Memory (U21, 62C1024Q) and EPROM (U18, 27C512) which is a bootstrap firmware for the 68HC11 (U25) microprocessor (U25). Sheet 4 (DPFC2) The ambient temperature reading circuit is realized through • TL084C operational amplifier (U37) • OP200GP operational amplifier • IC AD7715 (U38). The +15 Vdc is generated by the 78L05 voltage regulator (Q15) supplied to +5 Vdc. There are three identical circuits (U42, U41, U40, U39, U44, U43) to provide: • Three analogical set-point signals (10 mV/°K; 0V @-273 °C; 6.730 V @ 400°C) for three temperature control zones (OVEN, INJECTOR, DET BASE BODY). The drawing of Figure 4-12 shows the principle of the thermoregulation of three temperature zones controlled by microprocessor technologies. 110 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description MY04 Main Board Sections Description Figure 4-12. Thermoregulation Principle Sheet 5 (Power Supply; Valves; Stepper Drivers) This is the power supply section which provides the following voltages to all other blocks: • +5 Vdc @ 800 mA (CPU; DPFC; TEMP.CNTR; REF; KEYB; DISPLAY) • +15 Vdc @ 400 mA (DPFC) • -15 Vdc @ 100 mA (DPFC) • +12 Vdc @ 4.5 W (FLAP STEPPING MOTORS). • +24 Vdc @ 4 W (FID DET) • +8 Vdc @ 16 W (FID COIL IGNITER). There are three transformers (TR1, TR2, TR3) and three diode components (D1,D2, D3) before the six voltage regulators (Q4, Q12, Q5, Q13, Q6, Q14) to generate all above voltages which are “under voltage protection” control by CE9250 (U10; hybrid circuit). According to the GC configuration, the primary of the dedicated thoroidal transformer for the FID or TCD supply is connected through P11 connector (pins 4, 5, 9, 10). Only in the case of FID, the secondary of the thoroidal transformer is also connected through P11 connector (pins 1, 6). Thermo Scientific FOCUS GC - Service Manual 111 Chapter 4 Electronic Boards Description MY04 Main Board Sections Description The stepping motor (four phases, 3600 steps per revolution) for the modulating oven flap system is supplied by IC ULN2003A driver (U74). The +5Vdc is monitored by LED I1. Sheet 6 (Oven; Zone Heaters; Drivers) There are four SSR 6001 solid-state relay circuits (SSR1, SSR2, SSR3, SSR4) and three triacs (Q1 BTA26-600B and Q2, Q3 BTA06-600B) to control three heaters: • OVEN HEATER (1000 W) • INJECTOR (100 W) • DET BASE BODY (100 W) • one OVEN FAN MOTOR (25 W) The main board can be configured for 230 Vac or 120 Vac main input voltage by placing a different position of the connectors P3 and P2. Two F-2A fuses (F1, F3) protect the OVEN HEATER and DET HEATER, the fuses F-1.25A (F4) and F-1.25A (F5) protect the OVEN FAN MOTOR, while the fuse T-1.6A (F2) protects the transformer input. Two relays MZP-A001-4716 (K1, K2) switch off the main power voltage (230 or 120 Vac) to the heaters and fan motor when a “safety cut off” (alarm status) is activated by SCOFR safety circuit with BC337 (Q9) and diodes (D14, D15). The main inlet power voltage (230 or 120Vac) enter in the P1 connector (N, L, EARTH) by 10A POWER BREAKER and MAIN INPUT SOCKET with internal filter. Sheet 7 (Oven; Zone Thermoregulators) There are three identical circuits to control the reading temperature (ACTUAL TEMP VALUE) of three zones (OVEN, INJECTOR, DET BASE BODY) realized through: 112 • Three temperature sensors inputs (P8, P9, P10-called PT100 or RTD) • Three ICs CE9903 (U1, U2, U3 hybrid circuits) • Three ICs CE9902 (U4,U5, U6) Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description ELI201 Electrometric Board (FID) The switch of the OVEN DOOR is connected to P19 connector, then the circuit with R83 and BAS16 diode (D11) give the status of the oven door (open or close) to able or disable the driving of heaters and motors. ELI201 Electrometric Board (FID) This paragraph provides the description of the ELI201 electrometer board. Schematics and Component Layout The schematics and component layout information for the ELI201 board is listed below in Table 4-12. The number of drawings has been included for each schematic and its corresponding layout. Table 4-12. ELI210 Electrometer Board Schematics Board ELI201 PN Drawing Layout 236 485 25 SE 1467 (1/2) PC 1166 Board Description The main purpose of this board is to control the following parameters of the FID detector through a small electrometer pre-amplifier circuit marked CVA201. Refer to CVA201 Electrometer Pre-amplifier Board (FID). • IGNIT (flame igniter coil, about 2.0 A @ 8 Vac) • -300 V (flame jet polarization) • INPUT (low current produced) Electrometer board takes all digital inputs/outputs (SPI) and the power voltages (+5Vdc, ±15Vdc, +24Vdc) from the MY04 main board through P1 connector (32+32 pins, male). See also P30 connector on the MY04 main board. IMPORTANT The board requires the use of the assy cable FC 1399 having P/N 230 435 12. Refer to FC 1399 and SC 818. Thermo Scientific FOCUS GC - Service Manual 113 Chapter 4 Electronic Boards Description ELI201 Electrometric Board (FID) Sheet 1/2 (SP1; V/F; Power Supply) The serial communication lines (MOSI, MISO, CS, CLK) enter to IC 74HC240 (U16) which outputs (SCK, ACK, SIS) are isolated through three 6N137 (U17,U18,U19). The IC VFC110 (U21) converts the analogical voltage input (VIN, 0-10V) produced from FID detector through CVA pre-amplifier board to generate the output frequency (FOUT, from 0 to 3.5 MHz) which be read from microprocessor. The hybrid circuit CE9910 (U1) is supplied with ±15 Vdc to produce the –300 V output voltage for polarizing the FID detector. The igniter coil of FID detector is supplied to 8 Vac @ 2 A through a K1 relay which is driven from one BC327 (Q1) and BC546 (Q2) selected by two microprocessor signal (04, 05) of the IC 74HC595 (U6). The circuit realized by TL1021 (U26), hybrid circuit CE9557 (U22) and three voltage regulators LM7805 (U25), LM317 (U23) and LM337 (U24) is a stabilized power supply for ±15Vdc. SE 1467 Sheet 2/2 (Auto zero; Gain; Range) The amplifier OPA177 (U14) receives the input current from FID detector (INA, CVA) and the “auto zero” input current: • IOUT • RV1 • R32 Two ICs 74HC595 (U6, U9) receive the digital inputs from microprocessor (16 bits) to generate the auto zero current: • FINE by DAC08, U11 • GROSS by DAC08, U8). The circuit with ICs 74HC595 (U6), 74HC138 (U5) and LM324 (U15) select the range of electrometer (2 bits/4 range: 10°; 0,1,2,3,4) and the analogical filter (N3). 114 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description CVA201 Electrometer Pre-amplifier Board (FID) The amplifiers OP07 (U3), LM358 (U4) and DG413 (U10) select the “gain” of the frequency: • L = gain 1 • H = gain 32 The analogical outputs (1 V, 10 V) for external integrator are generated from OP177 (U13), C8, C9, R35, R36, R38 and R39. CVA201 Electrometer Pre-amplifier Board (FID) This paragraph provides the description of the electrometer pre-amplifier board. Schematics and Component Layout The schematics and component layout information for the CVA201 board is listed below in Table 4-13. The number of drawings has been included for each schematic and its corresponding layout. Table 4-13. CVA201 Pre-amplifier Board Schematics Board CVA201 PN Drawing Layout 236 485 70 SE 1462 (1/1) PC 1162 Board Description The “low current” signal (1 pA/10 mV; DET CABLE, SHIELD) produced from the FID detector is connected to the input board (G1) by a dedicated cable with coaxial connector, which goes to the DUAL-FET amplifier (U2) and operational amplifier OP07 (U1). Four ICs PN411 FET (Q1, Q2, Q3, Q4) are selected by the three “range” signals (S1, S2, S3; input current points G2) and N3 signal for analog filter through four 100 kΩ resistances and four BRY73 diodes. Then, the output voltage is available on pin 7 of P1 (VOUT). NOTE In order to verify the FID “electrometer noise level” it is necessary to make the following technical procedure: 1. Remove the coaxial connectors of the signal cable from FID detector Thermo Scientific FOCUS GC - Service Manual 115 Chapter 4 Electronic Boards Description TCDI203 Electrometric Board (TCD) 2. Wrap up the coaxial connectors using aluminum paper 3. Connect the analog output signal (0-1V) to the integrator 4. Set parameters ATT=0 and RNG=0 (max sensitivity) 5. The noise level must be within 1% of full scale TCDI203 Electrometric Board (TCD) This paragraph provides the description of the TCDI203 board. Schematics and Component Layout The schematics and component layout information for the TCDI203 board is listed below in Table 4-14. The number of drawings has been included for each schematic and its corresponding layout. Table 4-14. TCDI203 Electrometric Board Schematics Board TCDI203 PN Drawing Layout 236 487 65 SE 1522 (1/3) PC 1204 Board Description This board controls the SPI internal serial communication interface with MY04 main board, the analogical output signals (0V, 1V, 10V, GND), set functions and A/D converter. See P30 connector on the MY04 main board. Sheet 1/3 The integrated circuits U29A-B, U19, U20, U21 and U1A compose the section for SPI serial bus communication and optical-coupler interface with MY04 main board signals. Two ICs U30 and U1F is a circuit to read and control the PRESSURE SWITCH for the presence of carrier gas, while U1B-E are for internal buffer signals. The PLC U8 is a microprocessor controlled by internal clock of 1.8432MHz through OSC1. 116 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description TCDE203 Board (TCD) Sheet 2/3 In this section of board there four DAC circuits: U27/U28 for auto-zero, U5/U3 for set temperature filament, U4/U2 for set voltage and U17/U19A for set temperature cell. U6 with all resistors and capacitors is the ADC converter circuit for temperature cell. Sheet 2/3 The IC U11/U18B/U14/U15 is a circuit for GF (gain/frequency) selection of TCD detector, while the U12/U13/U16/U31A/U31B/U24 is a circuit for TCD analogical signal and offset. The U25 and U26 is a V/F converter circuit to generate a FOUT. TCDE203 Board (TCD) This paragraph provides the description of the TCDE203 board. Schematics and Component Layout The schematics and component layout information for the TCDE203 board is listed below in Table 4-15. The number of drawings has been included for each schematic and its corresponding layout. Table 4-15. TCDE203 Board Schematics Board TCDE203 PN Drawing Layout 236 487 70 SE 1523 (1/4) PC 1205 Board Description This board supplies the TCD cell and controls signals and temperatures. Sheet 1/4 The IC U5 and U2 constitute the TCD amplifier, while U6 circuit is for AUTOZERO. The U8A/B and U4 constitute the bridge sensing circuit. Thermo Scientific FOCUS GC - Service Manual 117 Chapter 4 Electronic Boards Description FRTMY100 Board and Keypad Sheet 2/4 The IC U13 and U15 is the bridge forcing circuit and supply, while U9A is for ON Filament circuit. U14 and Q2 are for sensing bridge amplifier. Sheet 3/4 The CR2, L2, U22 and Q3 is power supply circuit for +/-15VDC, while U24 and U25A is for +/- 10Vdc voltage. Sheet 4/4 The 48 Vdc is entering through F2 fuse and K1 relay to CR1 diode and U1 as power to heater. The PT100 temperature sensor is read from U2. U19, TE1 and U20 constitute the thermoregulation circuit, while U11, U7 and RV2 is for synchronize the main voltage. FRTMY100 Board and Keypad This paragraph provides the description of the Front Panel and the Keypad. Schematics and Component Layout The schematics and component layout information for the FRTMY 100 board is listed below in Table 4-16. The number of drawings has been included for each schematic and its corresponding layout. Table 4-16. FRTMY100 Board Schematics Board FRMTY100 PN Drawing Layout 236 485 16 SE 1465 (1/1) PC 1164 Board Description The front panel of the GC is very simple with few components and controlled by the MY04 main board through P28 and P33 connectors. 118 Thermo Scientific FOCUS GC - Service Manual Chapter 4 Electronic Boards Description IFM311 LAN Option There are: • Nine keys (from S1 to S9) • Four LEDS (I1, I2, I3, I4) • One LCD Display (2 lines x 20 characters) under control of P28 connector on MY04 main board IFM311 LAN Option The LAN option allows the control of the GC. This option is available only for GCs equipped with the MY04 Main Board. Schematics and Component Layout The schematics and component layout information for the IFM311 LAN option is listed below in Table 4-17. The number of drawings has been included for each schematic and its corresponding layout. Table 4-17. IFM3111 LAN Option Schematics Board LAN Option Thermo Scientific FOCUS GC - Service Manual PN Drawing Layout 432 102 15 SE 1527 (1/1) PC 1207 119 Chapter 4 Electronic Boards Description 120 IFM311 LAN Option Thermo Scientific FOCUS GC - Service Manual Maintenance and Troubleshooting Notes 5 This chapter contains the guides to maintain the S/SL injector and the FID detector, then analytical and technical troubleshooting notes. Chapter at a Glance… Maintenance .......................................................................................................122 Maintenance References.....................................................................................124 Analytical Troubleshooting Notes......................................................................126 Technical Troubleshooting Notes .......................................................................130 Thermo Scientific FOCUS GC - Service Manual 121 Chapter 5 Maintenance and Troubleshooting Notes Maintenance Maintenance This paragraph provides information on the maintenance of the instrument. It also contains the instructions for the instrument shutdown. Suggested Maintenance Schedule Preventive maintenance keeps your instruments running at peak performance. The recommended maintenance schedule is based on a GC running multiple samples in an 8-hour day. You may need to adapt the schedule to your situation depending on: • The number of analyses per day • How clean the samples are • The condition of the syringe needle • Environmental conditions such as dirt and ambient temperature Septa degrade over time. Different types will degrade at varying rates. Factors affecting septa life include: • Number of samples injected • Injector temperature • Condition of the syringe needle Our recommendations for septa replacement are based on 200 injections in a Thermo Fisher Scientific standard septum. Septa from other manufacturers may show different rates of degradation. Make sure you not only perform each scheduled item, but that you record it as well, along with any observation about instrument performance. Your records and notes can be invaluable when trying to trace a service problem. In addition to the routine and preventive maintenance that you perform, we recommend annual inspections by Thermo Fisher Scientific customer service engineers. These yearly checkups ensure that your instrument remains in peak condition. 122 Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Maintenance Standard Maintenance Schedule The following table reports the standard maintenance frequency. . Table 5-1. Standard Maintenance Schedule Frequency Daily Task Run Automatic Leak Test. Run performance check standard as required by method. Run calibration curve as required by method. Note gas pressure when the weekly pressure check shows it to be below 5000 kPa (50 bar or 750 psig). Weekly Change injector liner. Check pressure of gas cylinders. Replace the cylinder when its pressure drops below 3500 kPa (35 bar or 500 psi). Change injector septum. Perform a manual leak test on DPFC and Non-DGFC modules. Monthly Check for leaks from the primary gas supply to the GC. Semiannually Regenerate the injector inlet vent line trap. Replace FID jet. Annually Clean FID collector. Thermo Scientific FOCUS GC - Service Manual 123 Chapter 5 Maintenance and Troubleshooting Notes Maintenance References Maintenance References Maintaining a S/SL Injector The S/SL (Split/Splitless) injector will normally be serviced by Thermo Fisher Scientific authorized technical personnel. In order to operate at peak performances, the injector requires periodic maintenance from the user. This maintenance includes: • The replacement of the standard septum • The maintenance of the Merlin Microseal Valve when used instead of the standard septum • The cleaning or replacement of the liner • The replacement of the carbon filter from the DPFC control module. Refer to: • Chapter 15 of the FOCUS GC Instruction Manual Maintaining a PPKD Injector The Purged Packed Column injector and its interface module will normally be serviced by Thermo Fisher Scientific authorized technical personnel. In order to operate at peak performances, the injector requires periodic maintenance from the user. This maintenance includes: • The replacement of the septum • The cleaning or replacement of the liner. Refer to: • 124 Chapter 15 of the FOCUS GC Instruction Manual Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Maintenance References Maintaining a FID Detector To ensure optimum performance of the FID, you must keep it clean and free of dust and deposits. Symptoms such as reduced sensitivity and increased noise indicate that detector needs cleaning. To properly maintain the FID, you should perform the following cleaning or replacement sequences: • Clean/Replace the FID Jet • Clean/Replace the FID Collecting Electrode • Replace the FID Ignition Assembly Refer to: • Chapter 15 of the FOCUS GC Instruction Manual Maintaining a TCD Detector The TCD detector does not usually need current maintenance. Nevertheless, if you follow a few simple sequences, you will prolong the detector’s lifetime. Pay a special attention to avoid contamination or damage of the filaments. Follow these simple rules: • Avoid to activate/deactivate the bridge when not necessary. This operation could considerably reduce the filaments lifetime. • Avoid injecting samples that contain halogenated or acid compounds at high concentrations. • Ensure that oxygen (air) could not enter into the filaments cells. Oxidation would irreversibly damage the filaments. Install traps for moisture and oxygen on the gas lines to reduce the hazard. Refer to: • Chapter 15 of the FOCUS GC Instruction Manual Thermo Scientific FOCUS GC - Service Manual 125 Chapter 5 Maintenance and Troubleshooting Notes Analytical Troubleshooting Notes Analytical Troubleshooting Notes This paragraph contains a brief guide for the analytical troubleshooting. Behavior Characteristics Drifting Baseline Cause Remedy Accumulation of stationary phase Replace the end section of the column. Carrier gas cylinder pressure too low to allow control Replace the carrier gas cylinder or increase the pressure. Drifting carrier gas or combustion gas flows Check the gas controllers. Accumulation of impurities in column Check impurity levels in the gas source. Use correct gas purity. Carrier gas leak in the system Perform a leak test and ensure the tightness of the connections on the carrier gas line. Column is baking out Allow enough time for the column to stabilize. Accumulation of impurities in column Check impurity levels in the gas source. Use correct gas purity. Contaminated detector Check the detector and clean it. Column contaminated Recondition the column. Carrier gas flow rate too high Reduce the carrier gas flow. Contaminated column Recondition the column. Contaminated gases Replace gas cylinders or gas filters. Falling Rising Rising under temperature program operation High-standing current 126 Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Analytical Troubleshooting Notes Behavior Characteristics Cause Excessive column stationary phase bleeding High-standing current Irregular shape S-shaped Check the oven temperature, ensuring it doesn't exceed the column upper limit. Recondition the column. (continued) Dip after solvent peak Remedy Replace the column. Loose connections Ensure all interconnections and screw connections are tight. Detector contaminated Bake out or clean the detector. Excessive column bleed during column temperature programming Reduce the upper column temperature. Bake out the column. Install a high temperature column. Oxygen contamination is decomposing the stationary phase Install oxygen filters in carrier gas line. Check pneumatic and inlet systems for leaks. Use correct gas purity with low oxygen content. Square waves Noise High-frequency noise Large AC power fluctuations; heavy equipment on same line Use a dedicated clean AC line of sufficient amperage Contaminated detector Isolate the detector from electronics. If noise disappears, clean the collector. Combustion-gas flow too low or too high Check the detector-gas flows Contaminated column Condition the column. Contaminated detector gas supply Check the gas purity and install appropriate filters Defective electrometer Replace electrometer. Thermo Scientific FOCUS GC - Service Manual 127 Chapter 5 Maintenance and Troubleshooting Notes Behavior Noise (Continued) Characteristics High-frequency noise Analytical Troubleshooting Notes Cause Remedy Detector temperature higher than column maximum temperature Reduce the detector temperature to the column temperature upper limit. External electrical interference Attach an AC line monitor and check purity of the AC supply Loose column fittings Tighten fittings accordingly. Loose detector electrical connections Make sure the leads are properly connected. Defective electrometer or amplifier Replace the electrometer or amplifier F.s. column too close to flame (FID) Lower the column to the correct position (2-3 mm below the tip of the jet). Dirty jet or detector (FID) Isolate the detector from the electronics. If noise disappears, clean the jet and collector. External electrical interference Attach an AC line monitor and check the purity of the AC supply FID temperature too low Increase the temperature to at least 150 °C Operating conditions of the detector not properly set Optimize working parameters of the detector according to the actual operating mode (Constant Voltage, Constant Temperature). Spiking Refer to the FOCUS GC Instruction Manual for instructions. TCD Low Sensitivity Thermal conductivity of the carrier gas is too close to the one of the compound to be analyzed 128 Use a different carrier gas. Refer to the FOCUS GC Instruction Manual for instructions. Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Analytical Troubleshooting Notes Contaminated filaments TCD Low Sensitivity Remove contamination by baking the filaments for one hour at a temperature higher than the boiling point of the most highboiling compound. Perform baking twice if necessary. (continued) If the symptom does not disappears, contact customer support organization. Oxidized filaments Thermo Scientific FOCUS GC - Service Manual Contact your customer support organization. 129 Chapter 5 Maintenance and Troubleshooting Notes Technical Troubleshooting Notes Technical Troubleshooting Notes This paragraph contains the following technical troubleshooting notes: • Error Messages • Info/Diagnostics Error Messages The following Table 5-2 reports the error messages and the explanation of the relevant correlated cause and remedy: Table 5-2. Error Messages Message TEMP SHUTDOWN Loss of carrier Cause Remedy The inlet carrier supply is missing due to: 1. 2. Open the gas regulator Replace the cylinder 1. 2. Perform the injector leak test Replace the DPFC module The actual pressure cannot reaches the setpoint value due to: 1. 1. 2. 3. Inlet carrier pressure too low Possible injector leak DPFC module faulty 2. 3. Verify and adjust the correct inlet gas pressure Perform the injector leak test Replace the DPFC module The oven temperature exceeds the temperature limits. This error may be due to: 1. 2. 3. Replace the temperature sensor Replace the oven TRIAC Replace the MY04 Board 1. 2. Gas regulator close Cylinder empty The actual pressure value is 0 due to: 1. 2. HARDWARE SHUTDOWN Oven T over limits 1. 2. 3. 130 Possible injector leak DPFC module faulty Temperature sensor faulty (open) Oven TRIAC in short circuit MY04 Board faulty Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Technical Troubleshooting Notes Table 5-2. Error Messages (Continued) Message Cause HARDWARE SHUTDOWN Inlet T over limits The inlet temperature exceeds the temperature limits. This error may be due to: 1. 2. 3. HARDWARE SHUTDOWN Det T over limits 2. 3. 1. 2. 3. Replace the temperature sensor Replace the injector TRIAC Replace the MY04 Board 1. 2. 3. Replace the temperature sensor Replace the detector TRIAC Replace the MY04 Board Temperature sensor faulty (open) Injector TRIAC in short circuit MY04 Board faulty The detector temperature exceeds the temperature limits. This error may be due to: 1. Remedy Temperature sensor faulty (open) Detector TRIAC in short circuit MY04 Board faulty HARDWARE SHUTDOWN X line T over limits This message appears when a MS detector is connected to the GC instead of FID. The MS detector transfer line detector temperature exceeds the temperature limits. Refer to the relevant MS detector manual. TEMP SHUTDOWN Oven not heating The oven is not heating. This error may be due to: 1. 2. 3. 4. 5. 6. TEMP SHUTDOWN Inlet not heating Oven heater cable disconnected Oven door is open Oven door switch faulty Oven heater faulty Oven TRIAC open MY04 Board faulty The inlet not heating. This error may be due to: 1. 2. 3. 4. 5. Injector heater cable disconnected Injector heater faulty Injector TRIAC open Injector protection fuses MY04 Board faulty Thermo Scientific FOCUS GC - Service Manual 1. 2. 3. 4. 5. 6. Connect the oven heater cable Close the oven door Replace the oven door switch Replace the oven heater Replace the oven TRIAC Replace the MY04 Board 1. 2. 3. 4. 5. Connect the injector heater cable Replace the injector heater Replace the injector TRIAC Replace the F1 and/or F3 fuses Replace the MY04 Board 131 Chapter 5 Maintenance and Troubleshooting Notes Technical Troubleshooting Notes Table 5-2. Error Messages (Continued) Message Cause TEMP SHUTDOWN Det not heating Remedy The detector is not heating. This error may be due to: 1. 1. 2. 3. 4. 5. 2. 3. 4. 5. Detector heater cable disconnected Detector heater faulty Detector TRIAC open Detector protection fuses MY04 Board faulty Connect the detector heater cable Replace the detector heater Replace the detector TRIAC Replace the F1 and/or F3 fuses Replace the MY04 Board TEMP SHUTDOWN X-line not heating The MD detector transfer line not heating. Refer to the relevant MS detector manual. TEMP SHUTDOWN Hydrogen alarm Hydrogen sensor has detected the hydrogen gas into the oven. This error may be due to: 1. 2. 3. HARDWARE SHUTDOWN Undervoltage alarm WARNING! Before starting any operation, verify that the system has automatically cuts off the gas supply and heating to the active zones, and sweeps the column oven with forced air ventilation Probably leaks in the column connections Column breakage Hydrogen sensor faulty 1. 2. 3. Verify the column connections Replace the column Check hydrogen sensor following the instructions reported in the relevant manual The system is not properly operating. This error may be due to: 1. 1. • Verify if the main power line is correct If yes, the MY04 Board must be replaced If not, verify the main power line +5 Vdc; ± 15 Vdc supply to electric circuits is too low. • Info/Diagnostics Use this feature to bring up hardware, software and diagnostic information. In Main Menu move the cursor until the Others page is selected, then press ENTER. Scroll to Info/Diagnostics then press ENTER. 132 Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Technical Troubleshooting Notes OTHERS Info/Diagnostics A menu appears as the following: INFO/DIAGNOSTICS Serial #: 12345678 FW ver: 2.00.31 DPFC ver: 0.27 BOOT ver: 1 B.0 Manufact.: 1/Jan/02 Total runs 1234567 Ambient press 103.0 Table 5-3. Info/Diagnostics (1) Parameter Description INFO/DIAGNOSTICS This line is the title bar Serial #: The number (eight figures) indicates the product code. FM ver: Indicates the version of the current firmware of the instrument DPFC ver: Indicates the version of the DPFC pneumatic module installed BOOT ver: Indicates the version of the current BOOT required to load the firmware of the instrument. See FW ver: Manufact.: Indicates the manufacturing date of your GC. Total runs The number (seven figures), indicates the performed analyses. Ambient press Indicates the ambient pressure expressed in kPa. By pressing the sequence of keys SCROLL UP and SCROLL DOWN three times, four additional normally hidden lines are displayed. Thermo Scientific FOCUS GC - Service Manual 133 Chapter 5 Maintenance and Troubleshooting Notes Technical Troubleshooting Notes OV 31.8 F= 27 S=off IN 41.8 F= 30 S=off F= 25 S=off DT 25.3 OVPL 10 °C 0.7% Table 5-4. Info/Diagnostics (2) Parameter Description INFO/DIAGNOSTICS This line is the title bar OV, F, S Where: • OV = Oven raw temperature from A/D converter • F = Current temperature value displayed • S = Temperature setpoint value IN, F, S Where: • IN = Injector raw temperature from A/D converter • F = Current temperature value displayed • S = Temperature setpoint value DT, F, S Where: • DT = Detector raw temperature from A/D converter • F = Current temperature value displayed • S = Temperature setpoint value OVPL, X °C, Y% Where: • OVPL = Oven Power Level • X °C = Oven raw temperature from the A/D converter • Y % = Percentage of the power required to maintain the oven temperature 134 Thermo Scientific FOCUS GC - Service Manual Chapter 5 Maintenance and Troubleshooting Notes Technical Troubleshooting Notes Hydrogen Sensor Setting The presence of the hydrogen sensor is selected in Carrier Configuration menu. By pressing the sequence of keys SCROLL UP and SCROLL DOWN three times, an additional normally hidden line is displayed. CONFIGURE CARRIER Gas type Helium Hydrogen sensor? Y Selecting Y (yes) the hydrogen gas selection is enabled. CONFIGURE CARRIER Gas type Hydrogen Hydrogen sensor? Y WARNING! If the hydrogen sensor is not properly installed on the GC, the hydrogen gas selection as carrier gas is not possible. Changing the GC Manufacturing Data WARNING! This operation must be performed only for service purpose. Please contact the Thermo Fisher Scientific, Milan International Customer Support Department for instructions. Thermo Scientific FOCUS GC - Service Manual 135 Chapter 5 Maintenance and Troubleshooting Notes 136 Technical Troubleshooting Notes Thermo Scientific FOCUS GC - Service Manual 6 Schematic Drawings Summary This chapter contains the list of the drawings. Chapter at a Glance… Drawings List .....................................................................................................138 Thermo Scientific FOCUS GC - Service Manual 137 Chapter 6 Schematic Drawings Summary Drawings List Drawings List Use this table for a listing of all schematics and component layouts. The drawing and layout numbers have been included for easy reference: Table 6-1. Drawings List PCB Drawing Layout FOCUS GC with FID Schematic Diagram SE 1529; Rev.A; 1/3 SC 817; 1/1 FOCUS GC with FID Wiring Diagram --- SC 818; Rev.A; 1/1 FOCUS GC with TCD Schematic Diagram SE 1528; 1/3 SC 819; 1/1 FOCUS GC with FID Wiring Diagram --- SC 820; Rev.A; 1/1 MY04 Main Board SE 1526; 1/7 PC 1206; Rev.A; 1/1 ELI201 Electrometric Board for FID SE 1467; Rev.C; 1/2 PC 1166; Rev.F; 1/1 CVA201 Pre-amplifier Box for FID SE 1462; 1/1 PC 1162; 1/1 TCDI203 Electrometer Board for TCD SE 1522; Rev.A; 1/3 PC 1204; Rev.C; 1/1 TCDE203 Box for TCD SE 1523; Rev.A; 1/4 PC 1205; Rev.B; 1/1 Carrier Module Board SE 1453; Rev.E; 1/3 PC 1152; Rev.F; 1/1 IFM311 LAN Option SE 1527; 1/1 PC 1207; Rev.A; 1/1 FRTMY100 Front Panel SE 1465; 1/1 PC 1164; Rev.B; 1/1 Keypad Keypad 885 105 503; 1/1 --- Display LCD Display Panel 305AD5 (1/3) --- IEPC 138 Description Thermo Scientific FOCUS GC - Service Manual Glossary This glossary lists and defines terms used in this guide. Included are abbreviations, acronyms, metric prefixes, and symbols. A A ampere ac alternating current ADC analog-to-digital converter B b bit B byte (8 b) baud rate data transmission speed in events per second C ºC Celsius CIP Carriage and Insurance Paid To cm centimeter CPU central processing unit (of a computer) CSE Customer Service Engineer D d depth DAC digital-to-analog converter dc direct current DGFC Digital Gas Flow Control DPFC Digital Pressure and Flow Control DS data system Thermo Scientific FOCUS GC - Service Manual 139 Glossary E EMC electromagnetic compatibility EPROM Erasable Programmable ROM ESD electrostatic discharge F ºF Fahrenheit f frequency FID Flame Ionization Detector FOB Free on Board ft foot G g gram GC gas chromatograph GND electrical ground H h height h hour harmonic distortion A high-frequency disturbance that appears as distortion of the fundamental sine wave HV high voltage Hz hertz (cycles per second) I IEC 140 International Electrotechnical Commission Thermo Scientific FOCUS GC - Service Manual Glossary impulse See transient in. inch I/O input/output K k kilo (103 or 1024) K Kelvin kg kilogram kPa kilopascal L l length L liter LAN Local Area Network lb pound LED light-emitting diode M m meter (or milli [10-3]) M mega (106) µ micro (10-6) min minute mL milliliter mm millimeter m/z mass-to-charge ratio N Thermo Scientific FOCUS GC - Service Manual 141 Glossary n nano (10-9) non-DGFC Manual Detector gases Flow Control O Ω ohm P p pico (10–12) Pa pascal PCB printed circuit board PN part number PPKD Purged Packed Column Injector psi pounds per square inch R RAM random access memory RF radio frequency ROM read-only memory RS-232 industry standard for serial communications S 142 s second sag See surge slow average A gradual, long-term change in average RMS voltage level, with typical durations greater than 2 s S/SL Split/splitless injector Thermo Scientific FOCUS GC - Service Manual Glossary surge A sudden change in average RMS voltage level, with typical duration between 50 µs and 2 s T TCD Thermal Conductivity Detector transient A brief voltage surge of up to several thousand volts, with a duration of less than 50 µs V V volt V ac volts, alternating current V dc volts, direct current VGA Video Graphics Array W w width W Watt Thermo Scientific FOCUS GC - Service Manual 143 Index A About This Manual 9 Conventions Used 10 Overview 9 Air Flap 54 Air Flap Spring 58 Air Flap Stepping Motor 55 Analytical Troubleshooting 126 Automation Internal 22 Autosampler Interface 22 Autosampler Interface 22 Autosamplers Interface 38 Autosamplers Interface 38 B Base Unit 21 Board CVA201 115 ELI201 113 FRTMY100 118 MY04 (CPU) 97 TCDE203 117 TCDI203 116 C Calibration DPFC Module 35 Capillary Detector Base Body Manual Leak Test 33 CB71 Detector Assembly 71 Cleaning 18 Cleaning Gas Lines 30 Communication with External Units 22 Connectors 98 How to Dismounting 106 Conventions Used 10 CVA201 Electrometer Preamplifier 115 CVA201 Pre-amplifier Box 88, 89 Interface 22 Data Systems Interface 22 Decontamination 18 Detector 70 Detector Heater 74 Devices Option 38 Diagnostics 132 Document Set Use 14 DPFC Module 60 Calibration 35 DPFC Module Calibration 35 E Electrometer Board 113 Electrometer Preamplifier CVA201 115 Electronic Boards 86, 91 Electronic Cassette 81, 82, 94 Electronic Parts Replacements 80 ELI201 Board 87 ELI201 Electrometer Board (FID) 113 Error Messages 130 Exterior cleaning 18 External Devices Autosamplers 22 External Units Communication 22 F FID Detector 29 Fittings 25 Requirements 25 Flame Ionization Detector Controller Component Layout 108, 113, 115, 116, 117, 118, 119 FOCUS GC Document Set 14 Internal View 28 Series Differences 28 FOCUS™ GC Series 28 Front Panel 85 FRTMY100 Board 118 D Data Systems Thermo Scientific FOCUS GC - Service Manual 145 Index G K Gas Requirements 24 Gas Line Purging 30 Gas Lines 25 Cleaning 30 Requirements 25 Gas Regulators 25 Requirements 25 Gases Use 16 Gases Purity Requirements 24 GC Oven 28 Keypad 118 Kits Option 38 H Hazardous material decontamination 18 spills 18 Hydrogen Sensor Use 18 Use 17 Hydrogen Sensor 18 Setting 135 I Info 132 Injector 65, 92 Instrument Automation 22 Classification 15 Decontamination 18 Markings 12 Symbols 12 Technical Features 15 Interface Autosampler 22 Autosamplers 38 Data System 22 Internal Automation 22 Internal View 28 Instrument Cleaning 18 146 L LAN Option 119 Leak Test Capillary Detector Base Body Manual 33 External FOCUS™GC 31 M Main Cassette 92 Main MY04 Board Connectors Dismounting 38 Maintaining a FID Detector 125 Maintaining a PPKD Injector 124 Maintaining a S/SL Injector 124 Maintaining a TCD Detector 125 Maintenance 121, 122 Schedule Table 123 Standard Schedule 123 Suggested Schedule 122 Manufacturing Data 135 Changing 135 Marking 12 Markings 12 Material Safety Data Sheets 18 Mechanical Parts Replacement 42 Merlin Valve Kit 38 MY04 Board 87 Description 108 Introduction 97 N Non-DGFC Module 62 Notes Analytical Troubleshooting 126 Maintenance 121 Power Supply 101 Troubleshooting 121 O Option Devices 38 Thermo Scientific FOCUS GC - Service Manual Index Kits 38 Oven 42, 92 Oven Cryo System 39 Oven Door Switch 52 Oven Fan 50 Oven Heater 47 Oven Hydrogen Sensor 38 Oven Motor 51 Oven Sensor 45 P Packed Column Installation 39 Parts Replacement 41 Plumbing Requirements 24 Pneumatic Modules 59 Power Quality 23 Power Supply 101 Preliminary Tests 30 Purging Gas Lines 30 R Receptacle Pod 107 Receptacle Shroud 107 Regulations 30 Replacement Air Flap Spring 58 Air Flap Stepping Motor 55 CB71 Detector Assembly 71 CVA201 Pre-amplifier Box 88, 89 Detector Heater 74 Detector Thermal Sensor 74 DPFC Module 60 Electronic Cassette 82 Electronic Parts 80 ELI201 Board 87 Front Panel 85 Mechanical Parts 42 MY04 Board 87 Non-DGFC Module 62 Oven Door Switch 52 Oven Fan 50 Oven Heater 47 Oven Motor 51 Oven Sensor 45 Thermo Scientific FOCUS GC - Service Manual Parts 41 S/SL Heater 69 S/SL Injector Assembly 65 S/SL Thermal Sensor 69 Replacement TCD Detector Assembly 76 Requirements Fittings 25 Gas 24 Gas Lines 25 Gas Purity 24 Gas Regulators 25 Plumbing 24 S S/SL Heater 69 S/SL Injector 29 S/SL Injector Assembly 65 S/SL Thermal Sensor 69 Safety Information 16 Use of Gases 16 safety symbols 10–?? Schematic Diagrams 96 Spare Parts Electronics 138 Symbols 12 T TCDE203 Board 117 TCDI203 Electrometric Board 116 TCDI203 Electrometric Board (TCD) 116 Technical Features 15 Technical Notes 27 Time Events 119 Troubleshooting 121, 126 U Use of Gases 16 Using Hydrogen 17 Using the Hydrogen Sensor 18 147 Index 148 Thermo Scientific FOCUS GC - Service Manual Operating Sequences Cleaning and Purging Gas Lines .......................................................................................................30 External GC Leak Test ......................................................................................................................31 Injector Manual Leak Test .................................................................................................................32 Detector Base Body Manual Leak Test .............................................................................................33 DPFC Module Calibration .................................................................................................................35 How to Remove the Top and Rear Covers ........................................................................................43 How to Replace Oven Sensor ............................................................................................................45 How to Replace Oven Heater ............................................................................................................47 How to Replace the Oven Fan ...........................................................................................................50 How to Replace the Oven Motor .......................................................................................................51 How to Replace the Oven Door Switch .............................................................................................52 How to Replace the Air Flap Stepping Motor ...................................................................................55 How to Replace Air Flap Spring ........................................................................................................58 How to Replace the DPFC Module ...................................................................................................60 How to Replace the Non-DGFC Module ..........................................................................................62 How to Replace the S/SL or PPKD Injector Assembly .....................................................................65 How to Replace the S/SL or PPKD Temperature Sensor and Heater ...............................................69 How to Replace the CB71 Detector Assembly .................................................................................. 71 How to Replace the FID Temperature Sensor and Heater .................................................................74 How to Replace the TCD Detector Assembly ...................................................................................76 How to Replace the Complete Electronic Main Cassette ..................................................................82 How to Replace the Front Panel ........................................................................................................85 How to Replace MY04 and Electrometric Boards ............................................................................ 87 How to Replace CVA201 Pre-amplifier Box ....................................................................................88 How to Replace TCDE203 Box .........................................................................................................89 How to Dismounting the Connectors .................................................................................................106 Figures Figure 2-1. Figure 2-2. Figure 3-1. Figure 3-2. Figure 3-3. Figure 3-4. Figure 3-5. Figure 3-6. Figure 3-7. Figure 3-8. Figure 3-9. Figure 3-10. Figure 3-11. Figure 3-12. Figure 3-13. Figure 3-14. Figure 3-15. Figure 3-16. Figure 3-17. Figure 3-18. Figure 3-19. Figure 3-20. Figure 3-21. Figure 3-22. Figure 3-23. Figure 3-24. Figure 3-25. Figure 3-26. Figure 3-27. Figure 3-28. Figure 3-29. Figure 3-30. Figure 3-31. Figure 3-32. Figure 3-33. Figure 3-34. Figure 3-35. Oven Internal View ......................................................................................................28 Kit Calibration Box Connections .................................................................................36 Top Aluminium Cover Removal (1) ............................................................................ 43 Rear Panel.....................................................................................................................44 Back Cover Removal (1) ..............................................................................................45 Oven Temperature Sensor (1) ......................................................................................46 Oven Temperature Sensor (2) ......................................................................................46 Back Cover Removal (2) ..............................................................................................47 Main Board Cables Connection (1)..............................................................................48 Diaphragm Screws........................................................................................................49 Correct Oven Fan Position ...........................................................................................50 Main Board Cables Connection (2)..............................................................................51 Oven Motor Nut Position .............................................................................................52 FOCUS GC Right Panel...............................................................................................53 Door Switch..................................................................................................................53 Air Flap System............................................................................................................54 FOCUS GC Left Panel .................................................................................................55 Stepping Motor Cable...................................................................................................56 Security Dowel .............................................................................................................56 Stepping Motor Screws ................................................................................................57 Security Dowel Position ...............................................................................................57 Air Flap Clips ...............................................................................................................58 Brass Bush ....................................................................................................................59 DPFC Module Connections.......................................................................................... 60 DPFC Module...............................................................................................................61 Non-DGFC Module for FID Inlet Connections ...........................................................62 Non-DGFC Module for FID Cables Connections........................................................63 Non-DGFC Module for TCD Inlet Connections..........................................................64 Non-DGFC Module for TCD Cables Connections ......................................................64 Insulating Material Cover.............................................................................................66 Injector Temperature Sensor and Heater Screw ...........................................................67 Nut Inside the Oven......................................................................................................67 Column Adapter ...........................................................................................................68 Back Cover Removal (3) ..............................................................................................69 Main Board Cables Connection (3)..............................................................................70 Detector Insulating Material Cover ..............................................................................71 3-Line Pre-assembled Plate Fixing Screw.................................................................... 72 Figure 3-36. Figure 3-37. Figure 3-38. Figure 3-39. Figure 3-40. Figure 3-41. Figure 3-42. Figure 3-43. Figure 3-44. Figure 3-45. Figure 3-46. Figure 3-47. Figure 3-48. Figure 3-49. Figure 3-50. Figure 3-51. Figure 4-1. Figure 4-2. Figure 4-3. Figure 4-4. Figure 4-5. Figure 4-6. Figure 4-7. Figure 4-8. Figure 4-9. Figure 4-10. Figure 4-11. Figure 4-12. Detector Base Body Fixing Nut ...................................................................................72 Heater-Temperature Sensor Fixing Nut .......................................................................73 Back Cover Removal (4) ..............................................................................................74 Main Board Cables Connection (3)..............................................................................75 TCD Detector Assembly ..............................................................................................76 “-Line Pre-assembled Plate Removal)..........................................................................77 Back Cover Removal (5) ..............................................................................................77 Main Board Cables Connection (4)..............................................................................78 TCD Detector Assembly Removal............................................................................... 78 FOCUS™ GC Left Panel .............................................................................................83 Back Covers and Electronic Main Cassette Screw.......................................................83 FID Cables....................................................................................................................84 Back Covers and Electronic Cassette Screw (2) ..........................................................85 Front Panel Internal View ............................................................................................86 MY04 Main Board .......................................................................................................88 TCDE203 Box ..............................................................................................................90 FOCUS™ GC with FID Main Board Block Diagram .................................................94 FOCUS™ GC with TCD Main Board Block Diagram................................................ 95 MY04 Main Board .......................................................................................................97 Front View of Timed Event Connector ........................................................................ 100 Front View of Generic Handshake Connector .............................................................101 Front View of Sampler Signal Connector .................................................................... 103 Autosampler Connector................................................................................................104 Computer Connector .................................................................................................... 104 Back Cover ................................................................................................................... 106 Receptacle Shroud ....................................................................................................... 107 Receptacle Pod ............................................................................................................. 107 Thermoregulation Principle..........................................................................................111 Tables Table 4-1. Table 4-2. Table 4-3. Table 4-4. Table 4-5. Table 4-6. Table 4-7. Table 4-8. Table 4-9. Table 4-10. Table 4-11. Table 4-12. Table 4-13. Table 4-14. Table 4-15. Table 4-16. Table 4-17. Table 5-1. Table 5-2. Table 5-3. Table 5-4. Table 6-1. Main Schematic Diagrams ...........................................................................................96 Fuses .............................................................................................................................98 Connectors....................................................................................................................98 Timed Event Connector................................................................................................ 100 Code-Phases ................................................................................................................. 101 Generic Handshake Connector .....................................................................................102 Autosampler Connector................................................................................................103 Autosampler Connector................................................................................................104 Computer Connector ....................................................................................................105 Other Connectors.......................................................................................................... 105 MY04 Main Board Schematics ....................................................................................108 ELI210 Electrometer Board Schematics ..................................................................... 113 CVA201 Pre-amplifier Board Schematics ..................................................................115 TCDI203 Electrometric Board Schematics .................................................................116 TCDE203 Board Schematics.......................................................................................117 FRTMY100 Board Schematics ...................................................................................118 IFM3111 LAN Option Schematics ............................................................................. 119 Standard Maintenance Schedule...................................................................................123 Error Messages ............................................................................................................. 130 Info/Diagnostics (1)...................................................................................................... 133 Info/Diagnostics (2)...................................................................................................... 134 Drawings List ............................................................................................................... 138