BS-240 Chemistry Analyzer Service Manual © 2016 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved. For this Operator's Manual, the issue date is 2016-04. BS-240 Operator's Manual i Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. , , , , BeneView, WATO, BeneHeart, are the trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational or editorial purposes. They are the property of their respective owners. ii BS-240 Service Manual Responsibility on the Manufacturer Party Contents of this manual are subject to change without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this manual. Mindray is responsible for the effects on safety, reliability and performance of this product, only if: all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel; the electrical installation of the relevant room complies with the applicable national and local requirements; and the product is used in accordance with the instructions for use. Warning It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or personal injury. Note This equipment must be operated by skilled/trained clinical professionals. BS-240 Service Manual iii Warranty THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel. This warranty shall not extend to: Malfunction or damage caused by improper use or man-made failure. Malfunction or damage caused by unstable or out-of-range power input. Malfunction or damage caused by force majeure such as fire and earthquake. Malfunction or damage caused by improper operation or repair by unqualified or unauthorized service people. Malfunction of the instrument or part whose serial number is not legible enough. Others not caused by instrument or part itself. Customer service department Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray Building, Keji 12th Road South, High-tech industrial park, Nanshan, Shenzhen 518057,P.R.China Website: www.mindray.com E-mail Address: service@mindray.com Tel: +86 755 81888998 Fax: +86 755 26582680 EC - Representative iv EC-Representative: Shanghai International Holding Corp. GmbH(Europe) Address: Eiffestraβe 80, 20537 Hamburg, Germany Tel: 0049-40-2513175 Fax: 0049-40-255726 BS-240 Service Manual Preface Preface This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Please read this manual thoroughly before using the product. Observance of this manual is a prerequisite for proper performance and correct operation, and it ensures patient and operator safety. All graphics including screens and printouts in this manual are for illustration purpose only and must not be used for any other purposes. The screens and printouts on the actual product should prevail. Product introduction BS-240 is a computer-controlled fully-automated chemistry analyzer, intended for quantitative determination of clinical chemistries in serum, plasma, urine, cerebrospinal fluid (CSF), and other human body fluids. It can fulfill auto dispensing, reaction, colorimetric measurement, process monitoring, and result calculation. It provides measurement of multiple biochemistries and ISE (ion-selective electrode) tests (Na+, K+ and Cl-), with maximum throughput up to 400 tests per hour. It is one of the necessary tools for laboratory automation. BS-240 Service Manual 1 Safety Information Safety information This chapter provides you with safety symbols used in this manual and their meanings, summarizes the safety hazards and operating precautions that should be considered seriously when the instrument is being operated, and lists the labels and silkscreens that have been applied to the instrument and their indications. BS-240 Service Manual 1 Safety symbols Safety symbols are used in this manual in order to remind you of the instructions necessary to operate the product safely and in accordance with its function and intended use. A safety symbol and text constitutes a warning as shown in the table below: Symbol 2 Text Description WARNING Read the statement following the symbol. The statement is alerting you to an operating hazard that can cause personal injury. BIOHAZARD Read the statement following the symbol. The statement is alerting you to a potentially biohazardous condition. CAUTION Read the statement following the symbol. The statement is alerting you to a possibility of system damage or unreliable results. NOTE Read the statement following the symbol. The statement is alerting you to information that requires your attention. BS-240 Service Manual Summary of hazards This section lists hazards of the instrument itself. The hazards of specific operation are included in the warning information of each operation task. Observe the following safety precautions when using the product. Ignoring any of them may lead to personal injury or equipment damage. WARNING If the product is used in a manner not specified by our company, the protection provided by the product may be impaired. Electric shock hazards WARNING When the MAIN POWER is turned on, users other than the servicing personnel authorized by our company must not open the rear cover or side cover. Spillage of reagent or sample on the product may cause equipment failure and even electric shock. Do not place sample and reagent on the product. In case of spillage, switch off the power immediately, remove the spillage and contact our Customer Service Department or your local distributor. Moving Parts Hazards WARNING Do not touch such moving parts as sample/reagent carousel, reaction carousel, probe, mixer, and cuvette wash station, when the system is in operation. Exercise caution while using the ISE module. Prevent your hair, legs or other parts of your body from being hurt by the driving parts. Do not put your fingers or hands into any open part when the system is in operation. Photometer lamp hazards WARNING Eye injury could occur from light emission from the photometer lamp. Do not stare into the lamp when the system is in operation. If you want to replace the photometer lamp, first switch off the MAIN POWER and then wait at least 10 minutes for the lamp to cool down before touching it. Do not touch the lamp before it cools down, or you may get burned. BS-240 Service Manual 3 Laser beam hazards WARNING Light emitted by the bar code reader may cause eye injury. Do not stare into the laser beam radiated from the bar code reader when the system is in operation. Sample, calibrator and control hazards BIOHAZARD Inappropriately handling samples, controls and calibrators may lead to biohazardous infection. Do not touch samples, controls, calibrators, mixtures, or waste with your bare hands. Wear gloves and lab coat and, if necessary, goggles. In case your skin contacts the sample, control or calibrator, follow the standard laboratory safety procedure and consult a doctor. The serum samples remaining in the electrodes may contain a great number of viruses. Wear gloves to prevent infection while operating around the electrodes. Reagent and wash solution hazards WARNING Reagents and concentrated wash solution are corrosive to human skins. Exercise caution when using reagents and concentrated wash solution. In case your skin or clothes contact them, wash them off with soap and clean water. If reagents or wash solution spills into your eyes, rinse them with much water and consult an oculist. Waste hazards BIOHAZARD Some substances contained in reagent, control, calibrator, concentrated wash solution, and waste are subject to regulations of contamination and disposal. Dispose of the waste in accordance with your local or national rule for biohazard waste disposal and consult the manufacturer or distributor of the reagents for details. Wear gloves and lab coat and, if necessary, goggles. System disposal hazards WARNING Materials of the analyzer are subject to contamination regulations. Dispose of the waste analyzer in accordance with your local or national rule for waste disposal. 4 BS-240 Service Manual Fire and explosion hazards WARNING Ethanol is flammable substance. Please exercise caution while using ethanol around the instrument in order to prevent fire and explosion. Removal of analyzer from use for repair or disposal WARNING When the analyzer is not in use, for example, in repair, transportation or disposal process, please clean and sterilize the parts that may cause biohazards(probe, mixer, etc.) and remind the person who handles the device of the related hazards. BS-240 Service Manual 5 Summary of precautions This section lists precautions to be understood during instrument operation. The precautions of specific operation are included in the warning information of each operation task. To use the product safely and efficiently, pay attention to the following operating precautions. Intended use WARNING The instrument is an automated chemistry analyzer for in vitro diagnostic use in clinical laboratories and designed for in vitro quantitative determination of clinical chemistries in serum, plasma, urine and cerebrospinal fluid samples. Please consult us before you use the instrument for other purposes. When drawing a clinical conclusion, please also refer to patients' clinical symptoms and other test results. Environment precautions CAUTION Please install and operate the system in an environment specified by this manual. Installing and operating the system in other environment may lead to unreliable results and even equipment damage. To relocate the system, please contact our Customer Service Department or your local distributor. Electromagnetic noise precautions CAUTION Electromagnetic noise may interfere with operations of the system. Do not install devices generating excessive electromagnetic noise around the system. Do not use such devices as radio transmitters in the room housing the system. Do not use other CRT displays around the system. Do not use other medical instruments around the system that may generate electromagnetic noise to interfere with their operations. Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g. mobile phones or radio transmitters), as these may interfere with the proper operation. 6 BS-240 Service Manual The electromagnetic environment should be evaluated prior to operation of the device. This device has been designed and tested to CISPR 11 Class A, and in a domestic environment may cause radio interference, in which case, you may need to take measures to mitigate the interference. NOTE It is the manufacturer's responsibility to provide equipment electromagnetic compatibility information to the customer or user. It is the user's responsibility to ensure that a compatible electromagnetic environment for the equipment can be maintained in order that it will perform as intended. Operating precautions CAUTION Take the clinical symptoms or other test results of the patient into considerations when making diagnosis based on the measuring results produced by the system. Operate the system strictly as instructed by this manual. Inappropriate use of the system may lead to unreliable test results or even equipment damage or personal injury. When using the system for the first time, run calibrations and QC tests to make sure the system is in proper state. Be sure to run QC tests every time when you use the system, otherwise the result may be unreliable. Do not uncover the reagent carousel when the system is in operation. Keep the reagent carousel cover closed. The RS-232 port on the analyzing unit is used for connection with the operation unit only. Do not use it for other connections. Use the cables provided by our company or your local distributor for the connection. The operation unit is a personal computer with the operating software installed. Installing other software or hardware on the computer may interfere with the system operation. Do not run other software when the system is working. Computer virus may destroy the operating software or test data. Do not use the computer for other purposes or connect it to the Internet. If the computer is infected by virus, please install anti-virus software to check for and clear virus. Do not touch the display, mouse or keyboard with wet hands or hands with chemicals. Do not place the MAIN POWER to ON again within 10 seconds after placing it to OFF; otherwise the system may enter the protection status. If it does so, place the MAIN POWER to OFF and place it to ON again. BS-240 Service Manual 7 Chemistry parameter configuration precautions CAUTION To define such parameters as sample volume, reagent volume and wavelength, follow the instructions in this manual and the instructions of reagents. ISE module precautions CAUTION To prevent ISE electrodes from being damaged due to water scarcity, if the system, when equipped with an ISE module, will be powered off for a long time, perform the electrode storage maintenance. Sample precautions CAUTION Use samples that are completely free of insoluble substances like fibrin or suspended matter; otherwise the sample probe may be clogged. Medicines, anticoagulants or preservative in the samples may lead to unreliable results. Hemolysis, icterus or lipemia in the samples may lead to unreliable test results; running a sample blank, therefore, is recommended. Store the samples properly. Improper storage may change the compositions of samples and lead to unreliable results. Sample volatilization may lead to unreliable results. Do not leave the sample open for a long period. Prepare sufficient sample volume before analysis. Load samples to correct positions on the sample carousel before the analysis begins; otherwise reliable results may not be obtained. 8 BS-240 Service Manual Reagent, calibrator and control precautions CAUTION Use proper reagents, calibrators and controls on the system. Select appropriate reagents according to the performance characteristics of the system. Consult the reagent suppliers, our company or our authorized distributor for details, if you are not sure about your reagent choice. Store and use the reagents, calibrators and controls strictly as instructed by the suppliers; otherwise, reliable results or best performance of the system may not be obtained. Improper storage of reagents, calibrators and controls may lead to unreliable results and bad performance of the system even in validity period. Perform calibration after changing the reagents, otherwise reliable results may not be obtained. Contamination caused by carryover among reagents may lead to unreliable test results. Consult the reagent suppliers for details. ISE calibration precautions BIOHAZARD The calibrators contain preservatives. In case your skin contacts calibrators, wash them off with soap and water. In case the calibrators spill into your eyes, rinse them with water and consult an oculist. If you swallow them by mistake, see a doctor. CAUTION Use the calibrators specified by our company. Use of other reagents or calibrators may result in unreliable results, or damage the Hydropneumatic system, or even shorten the electrodes life span. Prior to using the calibrators, check if they are within the expiration date. Place them correctly; otherwise, it may cause unreliable results, or leak, or module damage. ISE wash solution biohazards BIOHAZARD The ISE wash solution is sodium hypochlorite. Use the ISE wash solution carefully to prevent it from contacting your skins or eyes. If your skins or eyes contact the ISE wash solution, rinse them off with fresh water and consult a doctor. BS-240 Service Manual 9 Data archiving precautions NOTE The system automatically stores the data to the built-in hard disk. Data loss, however, is still possible due to mis-deletion or physical damage of the hard disk. You are recommended to regularly archive the data to such medium as CDs. To avoid the data loss caused by unexpected power failure, UPS (uninterrupted power supply) is recommended. External equipment precautions WARNING For operating instructions and precautions of the computer and printer, please refer to their operation manuals. External equipment connected to the analogue and digital interfaces must be authorized and complied with relevant safety and EMC standards (e.g., IEC 60950 Safety of Information Technology Equipment Standard and CISPR 22 EMC of Information Technology Equipment Standard (CLASS B)). Any person, who connects additional equipment to the signal input or output ports and configures an IVD system, is responsible for ensuring that the system works normally and complies with the safety and EMC requirements. If you have any questions, consult the technical services department of your local representative. Tube and liquid container precautions WARNING When the tube or the part that contain liquid become aged or damaged, please stop its use immediately and contact our customer service department or your local distributor to check and replace it. 10 BS-240 Service Manual Labels and silkscreen The following non-warning and warning labels and silkscreen are used on the product for system identification and operating instruction. Check the labels regularly for cleanliness and integrity. If any of the labels becomes vague or peels off, contact our Customer Service Department or your local distributor for replacement. Non-warning labels and silkscreen Serial number This symbol, contained in the product label which is attached to the rear cover of the system, indicates the production serial number of the product. Date of manufacture This symbol, contained in the product label which is attached to the rear cover of the system, indicates the manufacture date of the product. In vitro diagnostic equipment This symbol, contained in the product label which is attached to the rear cover of the system, indicates that the product is in vitro diagnostic equipment. European community representative This symbol, contained in the product label which is attached to the rear cover of the system, indicates the name and address of the authorized representative in the European Community. BS-240 Service Manual 11 WEEE label The following definition of the WEEE label applies to EU member states only. The use of this symbol indicates that this product should not be treated as household waste. By ensuring that this product is disposed of correctly, you will help prevent bringing potential negative consequences to the environment and human health. For more detailed information with regard to returning and recycling this product, please consult the distributor from whom you purchased the product. Main power switch: ON This symbol located on the main power switch indicates that the system power is on when the switch is toggled rightwards to expose this symbol and ON character and the green light is lightened. Main power switch: OFF This symbol located on the main power switch indicates that the system power is off when the switch is toggled leftwards to expose this symbol and OFF character and the green light is extinguished. All components including the reagent refrigeration system will be stopped. Analyzer power switch This symbol located on the analyzer power switch indicates that the analyzer power is on when the switch is on the dotted-circle portion and off when it is on the blank-circle portion. Network interface This symbol located on the network interface indicates the connection between the analyzer and the operation unit. 12 BS-240 Service Manual Serial port This symbol located on the serial port indicates the connection between the analyzer and the operation unit. Electrical ground This symbol indicates an electrical ground. Interfaces for fluid connection This symbol located on the right panel of the analyzer indicates the connection of fluid tubing. BS-240 Service Manual 13 Near ISE tube installation unit This symbol indicates a specific name of the ISE tube. Humidity limit This symbol indicates a humidity limit on a normally working instrument. Barometric limit This symbol indicates a barometric limit on a normally working instrument. Temperature limit This symbol indicates a temperature limit on a normally working instrument. 14 BS-240 Service Manual Warning Labels Biohazard warning This label indicating the risk of biohazardous infection is located in the following positions: Probe Waste outlet Waste tank Moving parts warning This symbol and text indicating the hazardous moving parts is located in the following positions: Probe Mixer Laser warning This symbol and text located near the bar code reader reminds you of not staring into the laser beam. Photometer lamp warning This symbol and text located on the lamp housing reminds you of not touching the lamp before it gets cool. BS-240 Service Manual 15 Probe collision warning This symbol and text located near the sample/reagent carousel and reaction carousel reminds you of not opening the cover to prevent from damaging the probe. Upper cover This symbol and text located on the transparent upper cover reminds you of not wipe the upper cover with ethanol or other organic solutions. ISE module This symbol and text located on the left side panel of the analyzer. Please turn off the main power before opening the small door. Liquid level floater This symbol and text is located near the DI water tank liquid level floater, Please do not take out the liquid level floater during test. Cuvette replacement window This symbol and text is located on the cuvette replacement window. 16 BS-240 Service Manual BS-240 Service Manual 17 18 BS-240 Service Manual Safety Information Table of Contents Intellectual Property Statement ............................................................................................................................................. ii Responsibility on the Manufacturer Party........................................................................................................................ iii Warranty ........................................................................................................................................................................................ iv Exemptions ..................................................................................................................................................................... iv Customer service department ................................................................................................................................ iv EC - Representative ..................................................................................................................................................... iv Preface ............................................................................................................................................................................................. 1 Product introduction .................................................................................................................................................... 1 Safety information ··································································································· 1 Safety symbols............................................................................................................................................................................... 2 Summary of hazards................................................................................................................................................................... 3 Summary of precautions .......................................................................................................................................................... 6 Labels and silkscreen .............................................................................................................................................................. 11 Non-warning labels and silkscreen..................................................................................................................... 11 Warning Labels................................................................................................................................................................ 15 1 System Description ······························································································1-1 1.1 Overview ............................................................................................................................................................................. 1-2 1.2 Components of Analyzing Unit ................................................................................................................................... 1-3 1.3 Functions of Analyzing Unit ........................................................................................................................................ 1-5 2 System Performance and Workflow ··········································································2-1 2.1 Major Specifications ....................................................................................................................................................... 2-2 2.1.1 General Specifications.................................................................................................................................. 2-2 2.1.2 Sample Specifications................................................................................................................................... 2-3 2.1.3 Reagent Specifications ................................................................................................................................ 2-5 2.1.4 Specifications of Reaction System .......................................................................................................... 2-6 BS-240 Service Manual I Table of Contents 2.1.5 Specifications of Operation Unit .............................................................................................................. 2-7 2.1.6 Typical Test Procedure ................................................................................................................................ 2-8 2.1.7 Startup Procedure .......................................................................................................................................... 2-8 2.1.8 Shutdown Procedure .................................................................................................................................... 2-9 2.1.9 Workflow Descriptions ................................................................................................................................ 2-9 2.1.10 Measuring Points ...................................................................................................................................... 2-10 3 Installation Requirements and Procedure ··································································3-1 3.1 Check before Installation ............................................................................................................................................... 3-2 3.1.1 Installation Environment ............................................................................................................................ 3-2 3.1.2 Configuration Check ..................................................................................................................................... 3-2 3.2 Installation Requirements ............................................................................................................................................. 3-3 3.2.1 Space and Accessibility Requirements for Installation ................................................................... 3-3 3.2.2 Power Supply and Noise ............................................................................................................................. 3-3 3.2.3 Water Supply and Drainage ...................................................................................................................... 3-4 3.3 Computer Requirements(for International Customers).................................................................................. 3-5 3.3.1 CPU ...................................................................................................................................................................... 3-5 3.3.2 Memory.............................................................................................................................................................. 3-5 3.3.3 Network Adapter............................................................................................................................................ 3-5 3.3.4 Serial Port ......................................................................................................................................................... 3-5 3.3.5 Hard disk defragment .................................................................................................................................. 3-5 3.3.6 Remove Screen Saver and System Standby ......................................................................................... 3-6 3.3.7 Screen Display Properties .......................................................................................................................... 3-7 3.4 Installation Procedure ................................................................................................................................................. 3-10 3.4.1 Unpacking ...................................................................................................................................................... 3-10 3.4.2 Installation ..................................................................................................................................................... 3-10 3.4.3 Fluidic Connection ...................................................................................................................................... 3-13 3.4.4 Connection of Analyzer ............................................................................................................................ 3-13 3.4.5 Startup ............................................................................................................................................................. 3-13 3.4.6 Fluidic Prime................................................................................................................................................. 3-15 3.4.7 Initial Maintenance Record .................................................................................................................... 3-16 3.4.8 Startup Initialization .................................................................................................................................. 3-17 3.4.9 Importing and Configuring Chemistry Parameters of Mindray Reagents ........................... 3-18 3.4.10 System Performance Test ..................................................................................................................... 3-21 3.5 Stop Use ............................................................................................................................................................................. 3-23 3.5.1 Stop Use for Short Term ........................................................................................................................... 3-23 3.5.2 Stop Use for Long Term ............................................................................................................................ 3-23 II BS-240 Service Manual Table of Contents 4 Software Introduction ··························································································4-1 4.1 Software Installation ....................................................................................................................................................... 4-2 4.1.1 Introduction of Installation Package ....................................................................................................... 4-2 4.1.2 Folder Structure ............................................................................................................................................... 4-2 4.1.3 Log Files ............................................................................................................................................................. 4-4 4.1.4 Normal Startup Procedure.......................................................................................................................... 4-5 4.2 Software Upgrading......................................................................................................................................................... 4-6 4.3 Software Uninstallation .................................................................................................................................................. 4-7 4.3.1 Uninstalling BS-240 Software ................................................................................................................... 4-7 4.3.2 Uninstalling SQL Database ......................................................................................................................... 4-7 4.4 Database Backup and Restoration ............................................................................................................................ 4-8 4.4.1 Database Backup ............................................................................................................................................ 4-8 4.4.2 Database Restoration ................................................................................................................................... 4-8 4.5 Demo Software Setup ..................................................................................................................................................... 4-9 5 Modules and Units ·······························································································5-1 5.1 Shell Assembly .................................................................................................................................................................. 5-2 5.1.1 Main functions ................................................................................................................................................ 5-2 5.1.2 Component Locations and FRU Details ................................................................................................. 5-2 5.1.3 Replacing sheilding cover ........................................................................................................................... 5-4 5.1.4 Removing and Reinstalling Left and Right Panels ............................................................................ 5-4 5.1.5 Removing and Reinstalling Front Panel ............................................................................................... 5-5 5.1.6 Removing and Reinstalling Panels .......................................................................................................... 5-6 5.1.7 Replacing Air Spring ..................................................................................................................................... 5-7 5.1.8 Cleaning Fans ................................................................................................................................................... 5-7 5.2 Frame Assembly ............................................................................................................................................................... 5-9 5.2.1 Main functions ................................................................................................................................................ 5-9 5.2.2 Component Locations and FRU Details .............................................................................................. 5-10 5.2.3 Replacing Front and Rear Dust Screens ............................................................................................ 5-10 5.3 Reaction Carousel Assembly .................................................................................................................................... 5-12 5.3.1 Main functions ............................................................................................................................................. 5-12 5.3.2 Locations and FRU details ........................................................................................................................ 5-13 5.3.3 Replacing Cables of Upper Cover Assembly ..................................................................................... 5-15 5.3.4 Replacing Heat Chamber Assembly Cables ...................................................................................... 5-16 5.3.5 Replacing Cables of Home Position Sensor and Coder Sensor ................................................. 5-19 5.3.6 Replacing Motor Assembly and Synchronous Belt ....................................................................... 5-21 BS-240 Service Manual III Table of Contents 5.3.7 Replacing Reaction Cuvette ..................................................................................................................... 5-23 5.4 Sample/reagent carousel assembly ...................................................................................................................... 5-24 5.4.1 Main functions ............................................................................................................................................. 5-24 5.4.2 Assembly Locations and FRU Details .................................................................................................. 5-25 5.4.3 Replacing Sample/Reagent Carousel Body Assembly ................................................................. 5-27 5.4.4 Replacing Temperature Sensor ............................................................................................................. 5-28 5.4.5 Replacing Peltier and Sealing Ring ...................................................................................................... 5-30 5.4.6 Replacing Cables of Home Position Sensor and Coder Sensor ................................................. 5-32 5.4.7 Replacing Sample/Reagent Carousel Motor Assembly ............................................................... 5-34 5.4.8 Replacing the synchronous belt ............................................................................................................ 5-35 5.4.9 Replacing Reagent Refrigeration and Radiating Fans.................................................................. 5-37 5.4.10 Replacing Bar Code Reader ................................................................................................................... 5-38 5.4.11 Replacing Reagent Carousel Antifogging Heating Assembly ................................................. 5-40 5.5 Sample Probe Movement Assembly ...................................................................................................................... 5-42 5.5.1 Main functions ............................................................................................................................................. 5-42 5.5.2 Assembly Locations and FRU Details .................................................................................................. 5-44 5.5.3 Replacing Horizontal Movement Photocoupler and Vertical Movement Photocoupler . 5-45 5.5.4 Replacing Sample Probe Assembly, Anti-collision Spring and Power Screw ..................... 5-46 5.5.5 Replacing Reagent Preheating Assembly .......................................................................................... 5-48 5.5.6 Replacing Level Sense Board ................................................................................................................. 5-49 5.5.7 Replacing Horizontal Rotating Synchronous Belt .......................................................................... 5-50 5.5.8 Replacing Horizontal Motor Assembly............................................................................................... 5-51 5.5.9 Replacing Sample Probe Drive Assembly .......................................................................................... 5-53 5.6 Mixer Movement Assembly ...................................................................................................................................... 5-55 5.6.1 Main functions ............................................................................................................................................. 5-55 5.6.2 Assembly Locations and FRU Details .................................................................................................. 5-56 5.6.3 Replacing Horizontal Movement Photocoupler and Vertical Movement Photocoupler . 5-57 5.6.4 Replacing horizontal rotating synchronous belt ............................................................................ 5-58 5.6.5 Replacing Horizontal Motor Assembly............................................................................................... 5-59 5.6.6 Replacing DC Motor, Mixer and Powder Screw ............................................................................... 5-61 5.6.7 Replacement of Mixer Drive Assembly .............................................................................................. 5-62 5.7 Wash Assembly .............................................................................................................................................................. 5-65 5.7.1 Main functions ............................................................................................................................................. 5-65 5.7.2 Locations and FRU Details ....................................................................................................................... 5-66 5.7.3 Replacing Wash Probe Assembly ......................................................................................................... 5-67 5.7.4 Replacing Wash Station Photocoupler ................................................................................................ 5-68 IV BS-240 Service Manual Table of Contents 5.7.5 Replacing Wash Probe Drive Assembly ............................................................................................. 5-69 5.7.6 Replacing Zero Position Photocoupler ............................................................................................... 5-71 5.8 Syringe Assembly.......................................................................................................................................................... 5-73 5.8.1 Main functions ............................................................................................................................................. 5-73 5.8.2 Assembly Locations and FRU Details .................................................................................................. 5-73 5.8.3 Replacing 2.5ml Syringe and 10ml Syringe ..................................................................................... 5-74 5.8.4 Replacing Syringe ....................................................................................................................................... 5-74 5.9 Photoelectric Unit .......................................................................................................................................................... 5-75 5.9.1 Functions and Parameters ...................................................................................................................... 5-75 5.9.2 Composition and Structure of Optical Assembly............................................................................ 5-76 5.9.3 Replacing Lamp ........................................................................................................................................... 5-78 5.9.4 Disassembling/Assembling Optical Assembly ............................................................................... 5-79 5.9.5 Replacing Filter ............................................................................................................................................ 5-81 5.9.6 Replacing Preamplifier Board ............................................................................................................... 5-82 5.9.7 Replacing lamp fan...................................................................................................................................... 5-83 5.9.8 Replacing Reaction Cuvette ..................................................................................................................... 5-84 5.9.9 Replacing AD collection board ............................................................................................................... 5-85 5.10 ISE Unit (Optional) .................................................................................................................................................... 5-87 5.10.1 Module Functions ..................................................................................................................................... 5-87 5.10.2 Structure and FRU List ........................................................................................................................... 5-87 5.10.3 Unclogging Waste Tubes ....................................................................................................................... 5-90 5.10.4 Replacing Pump Tube............................................................................................................................. 5-91 5.10.5 Replacing Calibrator Tube .................................................................................................................... 5-92 5.10.6 Inventory Calculation of Reagent Pack ............................................................................................ 5-93 6 Hardware Circuits ·······························································································6-1 6.1 Overview ............................................................................................................................................................................. 6-2 6.2 Hazards ................................................................................................................................................................................. 6-3 6.3 Summary of PCBAs ......................................................................................................................................................... 6-4 6.4 Functions of PCBA ........................................................................................................................................................... 6-7 6.4.1 Control Structure ............................................................................................................................................ 6-7 6.4.2 Main Control Board ....................................................................................................................................... 6-7 6.4.3 Power drive board ...................................................................................................................................... 6-12 6.4.4 AD collection board .................................................................................................................................... 6-16 6.4.5 Pre-amplification Board .......................................................................................................................... 6-19 6.4.6 Level sense board ....................................................................................................................................... 6-21 6.4.7 Reagent Refrigeration Board ................................................................................................................. 6-23 BS-240 Service Manual V Table of Contents 6.5 Power Supply System .................................................................................................................................................. 6-29 6.5.1 Power Supply System of Whole Unit................................................................................................... 6-29 6.5.2 Performance Indices ................................................................................................................................. 6-29 6.5.3 Power supply board .................................................................................................................................. 6-30 6.6 Wiring Diagram of Analyzer .................................................................................................................................... 6-35 7 Hydropneumatic System ·······················································································7-1 7.1 Overview ............................................................................................................................................................................. 7-2 7.2 Fluidic Diagram ................................................................................................................................................................. 7-3 7.3 Principles of Hydropneumatic System ................................................................................................................. 7-11 7.3.2 Probe wash module .................................................................................................................................... 7-11 7.3.3 Cuvette Wash Module ................................................................................................................................ 7-12 7.3.4 Inlet/outlet module ..................................................................................................................................... 7-13 7.3.5 Other ................................................................................................................................................................ 7-15 7.4 Introduction of Fluidic Actions................................................................................................................................. 7-17 7.4.1 Fluidic Initialization ................................................................................................................................... 7-17 7.4.2 Priming Wash Station ................................................................................................................................ 7-17 7.5 Removing and Reinstalling Hydropneumatic Components ......................................................................... 7-18 7.5.1 Overview ........................................................................................................................................................ 7-18 7.5.2 Probe and Mixer Wash Module.............................................................................................................. 7-19 7.5.3 Cuvette Wash Module ................................................................................................................................ 7-25 7.5.4 External Modules......................................................................................................................................... 7-31 8 Alignment··········································································································8-1 8.1 Basic Operation ................................................................................................................................................................. 8-2 8.1.1 Utility - Maintenance .................................................................................................................................... 8-2 8.1.2 Alignment fixtures ......................................................................................................................................... 8-3 8.1.3 Mechanical Reset ........................................................................................................................................... 8-4 8.2 Alignment Procedure...................................................................................................................................................... 8-5 8.3 Photoelectric Unit ............................................................................................................................................................. 8-6 8.3.1 Photoelectric collection position adjustment ...................................................................................... 8-6 8.3.2 Photoelectric gain adjustment .................................................................................................................. 8-8 8.4 Sample/reagent carousel unit ................................................................................................................................. 8-11 8.4.1 Reaction Carousel Rotary ........................................................................................................................ 8-11 8.4.2 Sample/Reagent Carousel Outer Ring Radial Position ................................................................ 8-13 8.4.3 Sample/Reagent Carousel Middle Ring Radial Position .............................................................. 8-14 8.4.4 Sample/Reagent Carousel Inner Ring Radial Position ................................................................ 8-15 VI BS-240 Service Manual Table of Contents 8.5 Probe unit ........................................................................................................................................................................ 8-16 8.5.1 Probe to Horizontal Position on Reaction Carousel ..................................................................... 8-16 8.5.2 Aligning Horizontal and Vertical Positions from Probe to Wash Well .................................. 8-17 8.5.3 Probe to Sample/Reagent Carousel Outer Ring ............................................................................. 8-18 8.5.4 Probe to Sample/Reagent Carousel Middle Ring ........................................................................... 8-18 8.5.5 Probe to Sample/Reagent Carousel Inner Ring ............................................................................. 8-19 8.5.6 Sample Probe Cuvette Height ................................................................................................................. 8-19 8.5.7 Vertical Extreme Position from Probe to Reagent Bottle ............................................................ 8-19 8.5.8 Sample Probe Rotary to ISE (Optional) .............................................................................................. 8-20 8.5.9 Sample Probe Rotary to SIC Height (Optional) ............................................................................... 8-20 8.6 Mixer Unit ........................................................................................................................................................................ 8-22 8.6.1 Adjusting Horizontal Position from Mixer to Reaction Carousel ............................................ 8-22 8.6.2 Adjusting Vertical Position from Mixer to Reaction Carousel .................................................. 8-24 8.6.3 Adjusting Horizontal Position from Mixer to Wash Well............................................................ 8-24 8.7 Setting Parameters of Cuvette Wash Unit ........................................................................................................... 8-26 8.7.1 Setting Parameters of Best Alignment Position ............................................................................. 8-26 8.7.2 Adjust Horizontal Position from Wash Probe to Reaction Cuvette ........................................ 8-27 8.7.3 Cuvette Wash Station Height .................................................................................................................. 8-28 8.8 Bar Code Unit (Optional) ............................................................................................................................................ 8-30 8.8.1 Bar Code Sticking Requirements ........................................................................................................... 8-30 8.8.2 Bar code Unit................................................................................................................................................. 8-31 8.8.3 Adjusting Bar Code Reader Position .................................................................................................... 8-31 8.8.4 Bar Code Stability Test ............................................................................................................................... 8-33 8.9 Hydropneumatic Unit .................................................................................................................................................. 8-35 8.9.1 Preparations before Alignment ............................................................................................................ 8-35 8.9.2 Checking Floater and Valve Status ........................................................................................................ 8-36 8.9.3 Syringe Check ............................................................................................................................................... 8-36 8.9.4 Fluidic Prime................................................................................................................................................. 8-37 8.10 Pyrology Unit ................................................................................................................................................................ 8-39 8.10.1 Sensor Configuration window ............................................................................................................ 8-39 8.10.2 Observe Temperature Curve ............................................................................................................... 8-40 8.10.3 Wash Solution Temperature Control ................................................................................................ 8-41 8.10.4 Reagent Refrigeration Alignment ..................................................................................................... 8-42 9 Proactive Maintenance ·························································································9-1 9.1 Overview ............................................................................................................................................................................. 9-2 9.2 Maintenance Tools ........................................................................................................................................................... 9-3 BS-240 Service Manual VII Table of Contents 9.3 Maintenance Procedure ................................................................................................................................................ 9-4 9.3.1 Execution order of maintenance items ................................................................................................. 9-4 9.3.2 Steps .................................................................................................................................................................... 9-5 10 Troubleshooting ······························································································ 10-1 10.1 Error Codes and Troubleshooting ........................................................................................................................ 10-3 10.2 Error feedback sheet ............................................................................................................................................. 10-35 VIII BS-240 Service Manual 1 System Description BS-240 Chemistry Analyzer Service Manual 1-1 1 System Description 1.1 Overview 1.1 Overview The BS-240 is a fully automated and computer-controlled chemistry analyzer designed for the in vitro determination of clinical chemistries in serum, plasma, urine, and cerebrospinal fluid (CSF) samples. BS-240 consists of the analyzing unit and operation unit. Figure 1.1 BS-240 chemistry analyzer The throughput of the analyzer is at most 200 tests/hour; The throughput of the ISE module: Serum/plasma: 300 tests/hour (Na+, K+, Cl-). Urine: 198 tests/hour (Na+, K+, Cl-). Maximum throughput of biochemistry and ISE: 400 tests/hour. 1-2 BS-240 Service Manual 1 System Description 1.2 Components of Analyzing Unit 1.2 Components of Analyzing Unit The BS-240 Chemistry Analyzer consists of the analyzing unit (analyzer), operation unit (computer), and output unit (printer). The analyzing unit performs all operations of sample analysis, which include dispensing sample and reagent, mixing, reaction and measurement, washing of cuvettes, ISE analysis, etc. The operation unit (computer) is installed with the operating software to control the running and operation of the analyzers and to process the test data. The printer is used to print test results. The operation unit and the analyzing unit are independent structurally and communicate with each other through the serial port. The operation unit sends data and instructions to the analyzing unit and acquires the data and status information from the analyzing unit. Figure 1.2 System structure Analyzer PC (Operating software) Chemistry Analyzer Printer BS-240 has a throughput of at most 200 tests per hour for single- or double- reagent analysis, and its maximum throughput is 400 tests per hour with ISE module. The working period is 18 seconds. The instrument is composed of the following components: one reaction carousel, one sample/reagent carousel, one sample probe, one mixer, one cuvette wash station, one optical measurement system, and one ISE module(optional). The optical measurement system, which is composed of "filters", performs photometric measurement to the reaction cuvettes that hold "sample+reagent" mixture. After testing, a phase-8 automatic cuvette washing is conducted by the automatic cuvette wash station. BS-240 Service Manual 1-3 1 System Description 1.2 Components of Analyzing Unit Figure 1.3 Layout of the analyzer's operation panel (1) (5) (4) (2) (1) Sample probe (3) Mixer (5) Cuvette wash station 1-4 (3) (2) Sample/reagent carousel (4) Reaction carousel BS-240 Service Manual 1 System Description 1.3 Functions of Analyzing Unit 1.3 Functions of Analyzing Unit The working procedure of the system is described as follows. 1 The system resets to initialize all mechanical units and wash the exterior and interior of the probes and mixers. 2 Perform 8-phase automatic washing for each reaction cuvette, and after phase 6, water blank is analyzed automatically. 3 Rotate the reagent carousel to the first specified reagent position, and lower the sample probe to aspirate the reagent. 4 The reaction cuvettes are carried to the reagent dispense position after the 8-phase washing. Then rotate sample probe with R1 added to the reaction carousel and dispense R1 into the reaction cuvette. 5 R1 is incubated for 10 periods in the cuvettes. 6 Rotate the sample carousel to the specified sample addition position, and lower the sample probe into the sample carousel to aspirate a specified amount of sample. 7 Rotate the reaction cuvette with R1 added to the sample addition position, and rotate the sample probe with sample aspirated to the reaction carousel and dispense samples into the reaction cuvette. 8 Rotate the reaction cuvette with samples added to the mixing position for mixing. 9 For double-reagent tests, rotate the reagent carousel to the second specified reagent position after a specific period, and lower the sample probe into the reagent carousel to aspirate reagents. 10 The reaction cuvettes are carried to the second reagent dispense position. And the sample probe with R2 added rotates to the reaction carousel and dispenses R2 into the reaction cuvette. 11 With R2 added, the reaction cuvette rotates to the mixing position for stirring. 12 During each period, the reaction cuvette receives photometric measurement (absorbance reading taking) when passing by the photometric unit. 13 The reaction cuvettes in which reaction is finished will be washed automatically when passing by the wash station. Table 1.1 Functions of system units Unit Name Probe unit BS-240 Service Manual Functions Performs aspiration and dispensing for all biochemistries and ISE tests. Performs aspiration and dispensing of reagents for all biochemistry tests. 1-5 1 System Description 1.3 Functions of Analyzing Unit Sample/reagent carousel unit The outer ring can hold 40 sample containers. The middle ring supports 40ml and 20ml reagent bottles, and can hold 40 sample containers via an adapter. The inner ring supports 20ml reagent bottles. A refrigerating environment which is constant within 2°C-12°C for 24 hours a day is provided. The outer and middle rings support bar code scanning. Reaction Unit 40 positions are available for plastic cuvettes. Carousel Mixer Unit Mixing reactant Photometric unit Forward splitting, filter, implementing 8-wavelength measurement, 340, 405, 450, 510, 546, 578, 630, 670nm Cuvette Wash Unit 8-phase automatic washing of reaction cuvettes, which is performed 2 times for 4 stations ISE Unit (Optional) Provides the function of ISE measurement 1-6 BS-240 Service Manual 2 System Performance and Workflow BS-240 Chemistry Analyzer Service Manual 2-1 2 System Performance and Workflow 2.1 Major Specifications 2.1 Major Specifications 2.1.1 General Specifications System Random selection, multi-channel, multi-chemistry, capable to pause test and adding new tests System structure Analyzing unit + computer + printer Sample type Serum, urine, plasma, CSF(cerebrospinal fluid), full blood, etc. Maximum number of tests run simultaneously 39 double-reagent tests/78 single-reagent tests Throughput At most 200 tests/hour (single/double-reagent), and at most 400 tests/hour (with ISE module) Reaction types Endpoint, Kinetic and Fixed-time; supporting single-/double-reagent tests, and single/double-wavelength tests. Reaction time The longest reaction time is 20 minutes for single-reagent test, and 10 minutes for double-reagent test. Reaction temperature 37±0.3°C. Test type Clinical chemistries, Immunoturbidimetry, Therapeutic Drug Monitoring (TDM). Predilution Dilution is conducted in reaction cuvette at the ratio of 3~125. 2-2 BS-240 Chemistry Analyzer Service Manual 2 System Performance and Workflow 2.1 Major Specifications Operation mode Chemistries are defined one by one via the operating software; panels and calculation tests are supported. Math Model Single-point linear, two-point linear, multi-point linear, Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, and Parabola. QC Westgard multi rules, cumulative sum check, and twin-plot. Result Printouts Capable of storing and outputting various types of data and charts, and calculating among different chemistries. Dimensions Analyzing unit: l×b×h: 690 mm×580 mm×595 mm; Weight Analyzing unit: 79 kg; STAT sample Emergent samples can be analyzed at any time with highest priority. Networking Allowed to connect with a LIS (Laboratory Information Management System). 2.1.2 Sample Specifications Sample loading Samples are loaded via sample carousel. Sample cuvette type Table 2.1 Sample tube volume and dead volume Sample Container BS-240 Service Manual Specification Sample cup(Beckman) Φ14×25mm, 0.5ml(Beckman microtube) Sample cup(Beckman) Φ14×25mm, 2ml(Beckman microtube) Sample cup(Beckman) Φ12×37mm, 2ml (Hitachi) Primary tube or plastic tube Φ12×68.5 mm Primary tube or plastic tube Φ12×99 mm Primary tube or plastic tube Φ12.7×75 mm 2-3 2 System Performance and Workflow 2.1 Major Specifications Sample Container Specification Primary tube or plastic tube Φ12.7×100 mm Primary tube or plastic tube Φ13 X 75 mm Primary tube or plastic tube Φ13 X 95 mm Primary tube or plastic tube Φ13 X 100 mm Least sample volume The same container requires a sample to be 8mm higher than the unacceptable sample level height. Sample carousel The sample and the reagent share a carousel, and all sample positions of the outer ring and the middle ring support bar code scanning. Sample positions on sample carousel The outer ring can hold 40 samples, the middle ring can hold at most 40 samples via an adapter, and the outer ring and the middle ring contain at most 80 sample positions in total. STAT sample Emergent samples can be analyzed at any time with highest priority. Sample volume 2μl ~ 45μl, with an increment of 0.1μl ISE: 70μl for serum and 140μl for urine Sample probe The sample and the reagent share one probe, featuring level detection (vertical collision detection) and level tracking. Sample probe cleaning The exterior and interior of the sample probe is washed with carryover rate less than or equal to 0.05%. Sample input mode (bar code) Name 2-4 Value Symbology Codabar, ITF, Code128, Code39, UPC/EAN, and Code93 Minimum bar code density 0.19mm~0.5mm Data bits 3-27 bits Format and content Defined by user Maximum width 55mm BS-240 Chemistry Analyzer Service Manual 2 System Performance and Workflow 2.1 Major Specifications Name Value Minimum height 10mm Maximum inclination angle ±5° Print quality No less than Class C according to the ANSI MH10.8M Print Quality Specification. Width narrowness and 2.5-3.0:1 2.1.3 Reagent Specifications Reagent loading Samples are loaded via reagent carousel. Reagent bar code Name Value Symbology Codabar, ITF, Code128, Code39, UPC/EAN, and Code93 Minimum bar code density 0.25mm~0.5mm Data bits 13~30 bits Format and content Defined by user Maximum width 55mm Minimum height 10mm Maximum inclination angle ±5° Print quality Class A (ANSI MH10.8M) Width and narrowness 2.5:1 Reagent refrigeration Reagent Refrigeration Temperature: 2~12°C Reagent addition approach Reagent is aspirated and dispensed precisely by syringe, which is capable of detecting fluid level and checking reagent volume. Supported reagent types R1, R2. Reagent Vol R1 volume 100μl~250μl, with an increment of 0.5μl; R2 volume 10μl~250μl, with an increment of 0.5μl. BS-240 Service Manual 2-5 2 System Performance and Workflow 2.1 Major Specifications Reagent carousel The sample and the reagent share a carousel, and the reagents can be set in the middle ring or the inner ring. The positions of middle ring reagents support built-in reagent bar code reader. Number and volume of reagent bottle The reagent positions support Mindray reagent bottles, which include: outer ring 40ml, middle ring 20ml, and inner ring 20ml. Reagent carousel reagent quantity The middle ring supports 40ml and 20ml reagent bottles, and the inner ring supports 20ml reagent bottles. When a radial position of the middle ring has been loaded with a 40ml reagent bottle, the inner ring is forbidden to be loaded with a 20ml reagent bottle. Reagent probe The sample and the reagent share one probe, featuring level detection (vertical collision detection) and level tracking. Reagent probe cleaning Both interior and exterior of the sample probe are washed. Prevention of reagent cross contamination Users are allowed to perform carryover settings, e.g. washing interior and exterior of reagent probes and inserting special wash between tests. 2.1.4 Specifications of Reaction System Optical pathlength of reaction cuvette Light pathlength 5mm Material of reaction cuvette 5mm×5mm×29.5mm, plastic cuvette; cuvette segment structure, each segment containing 5 cuvettes Number of reaction cuvettes 40 Mixing method Mix after the sample and R2 are added. Reaction volume 100~360μl 2-6 BS-240 Chemistry Analyzer Service Manual 2 System Performance and Workflow 2.1 Major Specifications Optical measurement method Forward splitting filter system, detecting each wavelength through a photodiode Wavelength 8 wavelengths: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm Light source 12V/20W tungsten-halogen lamp Wavelength accuracy ±2nm Photometric measurement method Measuring through a photodiode Number of wavelengths simultaneously measured for each test Supporting one or two wavelengths Absorbance range 0-4.0A, optical path: 10mm Resolution of photometer 0.0001OD 2.1.5 Specifications of Operation Unit Display monitor Screen resolution of 1280×1024 Operating system Supporting Window 8 (64 bit) Communication Interfaces RS232, compatible with TCP/IP. Printer Supporting three types of printer: inkjet printer, laser (black and white) printer and stylus printer. Data input Keyboard; mouse; bar code reader; LIS: HL7, ASTM1394 (communicating through the TCP/IP or serial interface of static IP address) BS-240 Service Manual 2-7 2 System Performance and Workflow 2.1 Major Specifications Data output Display monitor, printer, RMS (communicating through TCP/IP interface of static IP address), and LIS. Data storage Hard disk and USB interface. 2.1.6 Typical Test Procedure Mixer performs mixing Add sample to probe Add R1 to probe Add R2 to probe Mixer performs mixing Single Reagent Chemistries Available chemistries are finished Shutdown Procedure Wash Cuvette Double Reagent Chemistries Wash Cuvette Wait for steadiness of lamp and balance of reaction carousel temperature Complete absorbency measurement Startup Initialization Procedure Figure 2.1 Test flowchart Next test 2.1.7 Startup Procedure 1 The analyzer is powered on. 2 The operating software is run and starts checking the running environment. 3 Check the middle-low layer units status. 4 The software checks the parameters and status of the analyzer, and inquires the key parameters and status of all units. 5 If a bar code reader is configured, its handshake is performed. 6 Fluidic initialization is performed. 7 Mechanical reset is performed. 8 The reaction carousel temperature control is turned on. 9 The lamp is turned off and the dark current and maximum luminance value of the reaction carousel lamp are inquired. The lamp is turned on (it takes 5 minutes for the lamp to become steady. When 5 minutes are not reached and the system finishes its home procedure, the system enters incubation status). If the lamp has been turned on, neither turn off the lamp nor query the dark current. 10 System reset is executed. 11 Clean special exterior of the sample probe. 12 Turn on the cuvette cleaning water, preheat the wash solution, and preheat the reagent. 2-8 BS-240 Chemistry Analyzer Service Manual 2 System Performance and Workflow 2.1 Major Specifications 13 The reagent bar code is scanned (when the reagent bar code reader is configured.) 14 The ISE module stats initialization (only if an ISE module is configured.) 15 If the conditions for system standby are met, the system enters into Standby status. 2.1.8 Shutdown Procedure 1 When the system is in the Standby status, the shutdown procedure can be performed. 2 Turn off the lamp; turn off the reaction cuvette temperature control, cuvette cleaning water and wash solution preheating and reagent preheating. 3 Pumps and valves are powered off for resetting. 4 Execute MANT periods of the ISE for 10 times, and discharge the waste in the ISE tube into the reagent pack (only if the ISE module is configured). 5 The database is shut down. If the operation fails, the system gives a warning, and continues the shutdown after confirmation. 6 The system exits. 2.1.9 Workflow Descriptions Sequential actions of sample probe in each period (adding sample for biochemistry test and R1) Wash ->Rotate to sample carousel->lower to sample tube->aspirate sample->raise to “home” position vertically -> rotate to reaction carousel-> lower into the reaction carousel -> dispense sample-> raise to “home” position vertically -> rotate to wash well >wash->Rotate to reagent carousel->lower to reagent bottle->aspirate R1-> raise to “home” position vertically -> rotate to reaction carousel-> lower into the reaction carousel -> dispense R1 ->raise to “home” position vertically-> rotate to wash well. Sequential actions of sample probe in each period (adding sample for ISE test) Wash ->rotate to sample carousel->lower into the sample tube-> aspirate sample -> raise to “home” position vertically -> rotate to top of ISE sample injection port -> lower to ISE sample injection port -> dispense sample -> raise to “home” position vertically -> rotate to wash well -> wash -> go to next period BS-240 Service Manual 2-9 2 System Performance and Workflow 2.1 Major Specifications Sequential actions of sample probe in each period (adding sample for whole blood test) Wash ->rotate to sample carousel->lower to sample tube->aspirate sample->raise to “home” position vertically -> rotate to wash well->lower to special wash position->wash special outer wall-> raise to “home” position vertically -> rotate to reaction carousel-> dispense sample ->raise to “home” position vertically-> rotate to wash well->wash->lower to special wash position -> wash special outer wall->raise to “home” position vertically -> go to next period. Sequential actions of sample probe in each period (adding sample for R2 test) Wash ->rotate to reagent carousel->lower into the reagent bottle-> aspirate R2 -> raise to “home” position vertically -> rotate to reaction carousel -> lower into the reaction carousel -> dispense R2 -> raise to “home” position vertically-> rotate to wash well -> wash-> go to next period. Sequential actions of mixer in each period Rotate to wash well -> lower to wash well -> wash -> raise to wash well -> rotate to reaction carousel -> lower to reaction carousel -> mix -> raise to top of reaction carousel>go to next period. Sequential actions of reaction carousel 18 seconds per period. Depending on the need of the dispense action of the sample probe, stop 3 times after rotating 3 times, or stop 2 times after rotating 2 times. 2.1.10 Measuring Points Figure 2.2 Measuring points for single-/double-reagent test R1 T1 R1 incubates for 180s Set incubation S MixS time of R2 arbitrarily R2 MixR2 Maximum reaction time of double reagent is over 10 min Reaction end T11 Maximum reaction time of single reagent is over 20 min 2-10 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure BS-240 Chemistry Analyzer Service Manual 3-1 3 Installation Requirements and Procedure 3.1 Check before Installation 3.1 Check before Installation 3.1.1 Installation Environment Altitude:-400~2,000m The system is for indoor use only. The bearing platform (or ground) should be level (with gradient less than 1/200). The bearing platform (or ground) should be able to support at least 100Kg weight. The installation site should be well ventilated. The installation site should be free of dust. The installation side should not be in direct sun. The installation site should be kept away from a heat or draft source. The installation site should be free of corrosive gas and flammable gas. The bearing platform (or ground) should be free of vibration. The installation site should be kept away from large noise and power supply interference. Keep the system away from brush-type motors and electrical contact device that is frequently switched on and off. Do not use such devices as mobile phones and radio transmitter near the system. Ambient temperature: 15°C-30°C, with fluctuation less than 2°C/H during operation. Relative humidity: 35%-85%, without condensation. Provide air conditioning equipment if the room temperature does not meet the requirements. Install the equipment near a sewer for convenient discharging of waste liquid. Dimensions before packing: 690mm(length)×580mm(depth)×595mm(height) Dimensions after packing: 825mm(length)×695mm(depth)×795mm(height) 3.1.2 Configuration Check 3-2 BS-240 Chemistry Analyzer Packing list of accessory kit Packing list of ISE accessory kit (for ISE module only) BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.2 Installation Requirements 3.2 Installation Requirements 3.2.1 Space and Accessibility Requirements for Installation Figure 3.1 System Clearance Minimum 500 Wall Maximum 3000 Minimum 500 Minimum 500 708 Operation Unit Analyzer 1180 Front Unit: mm Minimum 500 3.2.2 Power Supply and Noise Power supply: 220V model: 220V-240V~50Hz, 220/230V~60Hz, 110/115V~ 60Hz; voltage fluctuation: ±10%. Line frequency: ±3Hz. Three-wire power cord with good grounding performance (ground voltage ≤ 5V). The instrument should be powered by a properly-grounded power supply no less than 3000VA, and the mains socket used to connect the instrument should have its load no less than 10A. Keep the instrument at least 2.5m away from the mains socket. If you are going to use a UPS to power the instrument, make sure that the UPS can provide power supply greater than or equal to 3000VA/2100W. WARNING Make sure the power socket is grounded correctly. Incorrect grounding may lead to electric shock or equipment damage. Measurement of ground voltage: Set the multimeter to the AC 250V scale, connect the black probe to the earth wire and the red probe to the live wire and neutral wire. If grounding is proper, the voltage between earth wire and neutral wire should be less than 5V, and the voltage between earth wire and live wire should be similar to that between live wire and neutral wire. Check if the power socket outputs voltage meeting the specified requirements and has a proper fuse installed. BS-240 Service Manual 3-3 3 Installation Requirements and Procedure 3.2 Installation Requirements 3.2.3 Water Supply and Drainage Water quality: meeting requirements of CLSI type II with specific resistance no less than 1 (MΩ.cm@25°C) and silicate less than 0.1mg/L. Water temperature: 5-32°C. Water is supplied to the instrument by using a deionized water tank that accompanies the instrument, with a tank volume of 15L. The deionized water tank should be placed on the ground. Make the instrument discharge low-concentration waste to the sewer, or to the lowconcentration waste tank. Low-concentration waste tube is shorter than 5m. Make sure that the low-concentration waste outlet is no less than 50mm wide. BIOHAZARD Dispose of liquid waste according to the local regulations. CAUTION The supplied water must meet the requirements of CLSI type II; otherwise, insufficiently purified water may result in misleading test results. 3-4 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.3 Computer Requirements(for International Customers) 3.3 Computer Requirements(for International Customers) 3.3.1 CPU At least Intel Dual Core, 2.6GHz 3.3.2 Memory At least 2GB for each RAM 3.3.3 Network Adapter The computer needs to be connected to an LIS or Internet network, and one network adapter (Intel 1000M network adapter) needs to be added. 3.3.4 Serial Port The computer should provide an RS232 serial port, which is used to connect it with the analyzer. 3.3.5 Hard disk defragment Install the operating system in the C drive and the operating software of the instrument in the D drive. Make sure that the C drive is over 30G and D drive over 100G, and the disk file system is of NTFS format. Deselect the two options at the bottom of the disk properties window: “Compress drive to save disk space” and “Allow Indexing Service to index this disk for fast file searching”. BS-240 Service Manual 3-5 3 Installation Requirements and Procedure 3.3 Computer Requirements(for International Customers) Figure 3.2 Hard disk defragment Operating System Microsoft Window 8 (64 bit) The default login user of the computer must belong to the Administrator group and has permissions for administrator. Application Software Except for the operating system, other application software must not be installed or reserved on the computer. If the anti-virus application has been installed, then remove the automatic scheduled scanning and add BS-240 (Workstation) and BSLOG to the trust list. 3.3.6 Remove Screen Saver and System Standby To remove the screen saver and system standby, select None in the Screen saver pulldown list box of the Screen Saver Settings window, and then select Change power settings -> Change plan settings to go to the Edit plan settings window. Perform the settings as shown in the figures below. 3-6 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.3 Computer Requirements(for International Customers) Figure 3.3 Remove screen saver Figure 3.4 Standby settings 1 Select None in the Turn off the display and Put the computer to sleep pull-down list boxes. Figure 3.5 Standby settings 2 3.3.7 Screen Display Properties Set the screen resolution as 1280*1024 pixels and color quality as Highest (32 bit). BS-240 Service Manual 3-7 3 Installation Requirements and Procedure 3.3 Computer Requirements(for International Customers) Disable Automatic Synchronization with Internet Time Server Double click the time icon on the lower-right corner of the task bar. The Date and Time Properties window is displayed. Figure 3.6 Disable automatic synchronization with Internet time server - 1 Select the Internet Time tab, and then deselect the checkbox in front of Synchronize with an Internet time server. Figure 3.7 Disable automatic synchronization with Internet time server - 2 Turn off automatic updates Turn off the automatic updates on the Windows Update window. 3-8 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.3 Computer Requirements(for International Customers) Figure 3.8 Turn off automatic updates BS-240 Service Manual 3-9 3 Installation Requirements and Procedure 3.4 Installation Procedure 3.4 Installation Procedure 3.4.1 Unpacking 1 Before unpacking, check the packing list and see if the following items are included: chemistry analyzer, accessory kit, deionized water tank, dilution wash solution tank, high-concentration waste tank, low-concentration waste tank, wash solution, and ISE module (accessory kit, optional). 2 The chemistry analyzer is about 100kg. Unload it from the truck with a forklift. 3 Use a utility knife and an adjustable wrench to unpack, and remove the bracket and screws. 4 Move the instrument into position. 3.4.2 Installation Taking out Sample Probe and Reagent Probes Remove the packaging material and plastics on the analyzer panels. Take out the sample probe from the accessory kit and insert them into the corresponding metal connectors. Note not to forget the white washers in the package. Figure 3.9 Install white washer 3-10 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.4 Installation Procedure Installing Probes Figure 3.10 Installation procedure of probes Washer Keep the washer steady Retaining screw and spring Earthing wire Fluid level sense board connector 1 Before installing the probe, remove the probe arm cover by loosen the screws on it. 2 Insert the probe downwards into the hole on the probe arm while aligning the screw hole on the probe plate to the rod inside the arm. 3 Install the washer in the tube connector. To replace the washer, remove the old one from the tube connector and install the new one. Connect the tube connector to the probe and then tighten it. 4 Fix the earthing wire of the probe to the earthing terminal inside the arm; connect the probe connector to the fluid level detection board. 5 Sleeve the spring on the rod and tighten the retaining screw. Pay attention to the spring direction and make the thread opening face downwards. 6 Pinch the probe by the part near the probe arm. Push the probe upwards and then release it to check if the spring works well. If it does, proceed to the next step. If not, check if the spring is clamped or fixed too tightly. 7 After installation, power on the chemistry analyzer, and then check if the No.D2 LED indicator on the circuit board inside the probe arm is lit. If it is, it indicates that the level detection system works normally. 8 BS-240 Service Manual Install the arm cover and tighten the screws on it. 3-11 3 Installation Requirements and Procedure 3.4 Installation Procedure 9 Pinch the probe by the part near the probe arm. Push the probe upwards and then release it to check if the spring works well. If it does, proceed to the next step. It not, it indicates that the arm cover is not installed correctly. Reinstall the arm cover and check the spring until it can move freely. Installing Mixers 1 Take out the mixer from the accessory kit and lift the mixer arm to the vertical extreme. Figure 3.11 Take out mixers 2 Screw the mixer into the large hole of the retaining screw till it is level with the small hole end. Figure 3.12 Mixer 3 Insert the mixer motor shaft into the mounting hole and make sure they contact tightly and there is no gap in the vertical direction. 4 Tighten the fixing screws of the mixer clockwise and check whether the mixer is perpendicular. If not, reinstall the mixer. Figure 3.13 Tighten the retaining nut and check if the mixer is vertical 3-12 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.4 Installation Procedure 3.4.3 Fluidic Connection 1 The deionized water tank is filled with deionized water. 2 Prepare the dilution wash solution at a proportion of 1:9 (1 portion of concentrated wash solution to 9 portions of deionized water), and fill the dilution wash solution tank with the dilution wash solution. 3 Use the deionized water tank connectors, dilution wash solution tank connectors, and waste tank connectors in the accessory kit to connect the tank to the deionized water inlet, the dilution wash solution inlet, the low-concentration waste outlet, and the high-concentrated waste outlet on the right side of the instrument. 3.4.4 Connection of Analyzer 1 If you are using an UPS, install it and make it properly grounded. 2 Connect the analyzer, computer, printer, display monitor, power cord and data cable. Do not switch on the power before finishing all connections. 3 Connect the computer with the analyzer with the serial port cable, and choose the COM1 port for the computer. 3.4.5 Startup 1 Turn on the main power switch and the analyzer power switch. 2 Start the computer. The BS-240.exe operating software is run automatically when the Windows is started. You are also allowed to run the operating software by selecting the shortcut icon on the desktop or in the Start menu. See the figure below. The system skips the initialization procedure when detecting that the fluidic prime is not performed. Enter the username “ServiceUser” and password “#BS8A#SEU” in the login window. BS-240 Service Manual 3-13 3 Installation Requirements and Procedure 3.4 Installation Procedure Figure 3.14 Startup screen Figure 3.15 "Login" Screen 3 When the main screen is shown, select Utility -> System Setup -> Factory. Select "Optional Modules", and enable the ISE module, sample bar code and reagent bar code, and then select Save. See the figure below: 3-14 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.4 Installation Procedure Figure 3.16 "Optional Modules" window 3.4.6 Fluidic Prime 1 Select Utility -> Maintenance -> Alignment -> Hydro Unit. The screen is displayed as follows. Figure 3.17 "Hydro Unit" screen 2 Select 3 Fluidic Prime to display the following window. The check of solenoid valve, level floater and pump and syringe has been finished by default. Click "Continue". Follow the screen instructions, remove the cuvette wash station and put it in an opening container to finish syringe priming, and then install the wash station back. BS-240 Service Manual 3-15 3 Installation Requirements and Procedure 3.4 Installation Procedure 3 Complete cleaning and priming of inner and outer walls of the sample probe and the mixer. 4 The fluidic prime is finished. 3.4.7 Initial Maintenance Record Select Utility -> Maintenance -> Maintenance to display the maintenance window. Confirm all scheduled maintenance procedures and set up start time for them. The maintenance frequencies include Daily, Weekly, Two-Week, Monthly, Three-Month, Six-Month, and Other. Select the Select All button on each maintenance frequency tab page, and then select OK. The current date and time appears in the Date Performed column of each maintenance procedure as shown in the figure below. 3-16 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.4 Installation Procedure Figure 3.18 Scheduled Maintenance screen 3.4.8 Startup Initialization When the fluidic prime is finished, exit the operating software and reboot the computer. The operating software is automatically run and starts initializing. When the login window is displayed, enter the user name: "Admin" and password "Admin", as shown below. Figure 3.19 Startup screen BS-240 Service Manual 3-17 3 Installation Requirements and Procedure 3.4 Installation Procedure Figure 3.20 "Login" Screen When the main screen is shown, select Reagent -> Reagent/Calibration, and then perform ISE reagent loading. 3.4.9 Importing and Configuring Chemistry Parameters of Mindray Reagents 1 3-18 Select Utility -> Chemistries -> Config, and select Options to display the Option window. Select Import to display the Import window. BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.4 Installation Procedure Figure 3.21 Enter the "Import" window 2 Select Load Default to display all Mindray reagent chemistries in the left column. Select Add All to add all chemistries to the right column. Select Exit to close the window. BS-240 Service Manual 3-19 3 Installation Requirements and Procedure 3.4 Installation Procedure Figure 3.22 "Import" window 3 All Mindray reagent chemistries are displayed in the "Available Chemistries" column. You can delete chemistries that will not be used in your laboratory. Figure 3.23 "Chemistry Configuration" screen 3-20 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.4 Installation Procedure 4 Select OK to complete the chemistry configuration. Figure 3.24 Complete chemistry configuration 3.4.10 System Performance Test Water Test 1 2 Perform the following checks before starting test: All components' mechanical movement is correct without abnormal alarms (of optics, temperature control, mechanics, communication, etc.). Probes are washed normally without hanging liquid. When lowering to cuvette bottom, the wash station is lifted The reaction curve of all samples has no periodical fluctuation or sharp jumps, and the absolute value of test results is no greater than 10. Chemistry Setup Set a single-reagent chemistry "Water", with parameters including: endpoint, positive reaction, primary wavelength 340nm, secondary wavelength 405nm, result unit U/L, decimal places of result 0.01, blank time (-9)-(-7), incubation time 0, reaction time (13)-(22), sample volume 10, reagent volume R1:100. 3 BS-240 Service Manual On the Calibration Setup window, select the chemistry "Water", specify the K factor math model with K factor 10000, select Save and exit the window. 3-21 3 Installation Requirements and Procedure 3.4 Installation Procedure 4 Load the reagent. 5 Preparation for Analysis Load deionized water as R1 (each more than 10ml). Place deionized water as 1 sample (each more than 2ml). Place concentrated wash solution in position 39 on the reagent carousel. NOTE: Wear gloves when loading the wash solution. 6 Starting Analysis Program 1 sample with the chemistry "Water", with 40 replicates. Start the test and observe the working status of the parts. If an alarm is given, troubleshoot the error. 3-22 BS-240 Chemistry Analyzer Service Manual 3 Installation Requirements and Procedure 3.5 Stop Use 3.5 Stop Use 3.5.1 Stop Use for Short Term When to do Stop use for short term (within 3 days) How to do 1 Select Utility -> Maintenance -> Alignment -> Fluidic Emptying Unit. 2 Empty the fluidic tubes according to the software prompts. 3 If the ISE module is configured, select Empty ISE Tube to empty the ISE tubes according to the software prompts. Remove the ISE electrodes and inject proper amount of Calibrator A into the Na, K and Cl electrodes with a pipettor, and use tape to seal the electrodes before storing them. As regards the reference electrode, insert a soft tube into it. 4 Exit the operating software, shut down the computer and power off the analyzer. 5 If the analyzer is to be stored in another place, disconnect the data cables among the analyzing unit, computer, monitor, and disconnect the tubes in the back of the analyzer following the reversed order in Section 3.4.3. 3.5.2 Stop Use for Long Term When to do Stop use for long term (more than 3 days or for several years) How to do 1 Select Utility -> Maintenance -> Alignment -> Fluidic Emptying Unit; 2 Empty the fluidic tubes according to the software prompts. 3 If the ISE module is configured, select Empty ISE Tube to empty the ISE tubes according to the software prompts. Remove the ISE electrodes and inject proper amount of Calibrator A into the Na, K and Cl electrodes with a pipettor, and use tape to seal the electrodes before storing them. As regards the reference electrode, insert a soft tube into it. 4 Exit the operating software, shut down the computer and power off the analyzer. 5 Follow the reversed order in 3.4.4 Connection of Analyzer to disconnect the analyzer and follow the reversed order in 3.4.3 to disconnect the tubes in the back of the analyzer. BS-240 Service Manual 3-23 3 Installation Requirements and Procedure 3.5 Stop Use 3-24 6 Follow the reversed order in 3.4.2 Installation to restore the analyzer, and place the probes and related parts into the accessory kit for storing. 7 Follow the reversed order of unpacking in 3.4.1 Unpacking to push the main analyzing unit into the packing box for packing. BS-240 Chemistry Analyzer Service Manual 4 Software Introduction BS-240 Chemistry Analyzer Service Manual 4-1 4 Software Introduction 4.1 Software Installation 4.1 Software Installation 4.1.1 Introduction of Installation Package The installation package of the operating software contains three folders: Setup, SetupGuide, and Thirdparty. Figure 4.1 Folders in operating software installation package Setup folder: includes the operating software installation file (setup.exe) and the KillBsLog tool that can end the Bslog service. Figure 4.2 Setup folder KillBslog: a service run automatically when the operating software is started, and used to record the logs of the software. When started, the software checks if the service is started, and if it is not started, the software gives an alarm indicating startup failure. When logging on the software, you are allowed to end this process, and this action will not be recorded in the logs. SetupGuide folder: includes the software installation and upgrading guide. Thirdparty folder: includes dotnetfx and SQLExp installation packages. The former contains the installation program of Microsoft .net Framework and is the running environment assembly of SQL;while the latter is database software. 4.1.2 Folder Structure The default installation path of the operating software is: D:\Mindray\Workstation. 4-2 BS-240 Chemistry Analyzer Service Manual 4 Software Introduction 4.1 Software Installation After installation, the folder structure is like the following: Figure 4.3 Workstation folder The OperationSoft folder is relatively important and expanded as follows: Figure 4.4 OperationSoft folder Important folders include: AbsData: stores reaction curve files. AlarmFile: stores alarm log messages. Database: database file, which contains all parameters, test results, calibration and QC data, etc. ExceptionLog: records address segment for error analysis when the software collapses abnormally. (Note: There are two ExceptLog files, and the other one is in Instrument_1 folder.) BS-240 Service Manual Help: online help in .chm format. 4-3 4 Software Introduction 4.1 Software Installation Instrument_1: log file of analyzer. Item: Mindray's chemistry parameter file. LOG: records software upgrading information. (last upgrading information record) Logs: RMS (Remote Management System) logs. Not used temporarily. Workstation.exe: shortcut of the BS-240 program. 4.1.3 Log Files The BsLog process records in real-time mode the communication and action information of the software and analyzer, which includes but not limited to the following: Keyboard input Error log, operation log and maintenance log of the software Communication instructions between the PC and analyzer Action instructions of the analyzer Floater and photocoupler statuses of the analyzer (real-time) Pump and valve powering statuses of the analyzer (real-time) Photoelectric data (real-time) …… Separate log files are produced every day and stored in the following addresses: 1 WorkstationLOG folder at D:\Mindray\Workstation\OperationSoft. 2 InstrumentLOG folder at D:\Mindray\Workstation\OperationSoft\Instrument_1. The folder structure is shown below: 4-4 BS-240 Chemistry Analyzer Service Manual 4 Software Introduction 4.1 Software Installation Figure 4.5 InstrumentLOG folder Important folders include: InstrService: water blank data. Lislog: LIS communication data. RPC: (short form of Remote Procedure Call) basic control command data. ISEAdapterSrv: ISE command and test data. Workdata: instrument working instructions record 4.1.4 Normal Startup Procedure The normal startup procedure of the operating software is as follows: 1 The system checks if the Bslog process has been started automatically. If not, the Bslog.exe program is run. 2 Sqlserver is started, and it will exit if an error indicating database initialization failed occurs. 3 The software process is started. 4 Enter the username and password. 5 The software shakes hands with the middle-/lower-layer units. If handshake fails, an alarm is given prompting the unit is abnormal. BS-240 Service Manual 4-5 4 Software Introduction 4.2 Software Upgrading 4.2 Software Upgrading Refer to the upgrading guide in the SetupGuide folder or in the D:\Mindray\Workstation\UpgradeTool path. Note: After upgrading the software, switch off the main power and switch it on again to refresh the refrigeration unit. 4-6 BS-240 Chemistry Analyzer Service Manual 4 Software Introduction 4.3 Software Uninstallation 4.3 Software Uninstallation 4.3.1 Uninstalling BS-240 Software 1 End the BSlog process by using the KillBslog tool in the installation package. 2 Find the WorkStation program on the Control Panel window and uninstall it. 4.3.2 Uninstalling SQL Database Find the SQL Server program on the Control Panel window and uninstall it. BS-240 Service Manual 4-7 4 Software Introduction 4.4 Database Backup and Restoration 4.4 Database Backup and Restoration 4.4.1 Database Backup To back up a database, exit the operation software first, and then copy the BA80.bak file at D:\Mindray\Workstation\OperationSoft\ Database\Backup to a database file backup location of a local disk. 4.4.2 Database Restoration To restore a database, find a location of the to-be-restored database in the local disk first, and then copy the database file to D:\Mindray\Workstation\OperationSoft\ Database\Backup. 4-8 BS-240 Chemistry Analyzer Service Manual 4 Software Introduction 4.5 Demo Software Setup 4.5 Demo Software Setup To set the operation software to a Demo mode, exit the operation software first, and then add a command line in the WorkStation.exe startup program. Specifically, change the name WorkStation of the WorkStation.exe startup program to WorkStation -Demo MCUDemo –ISEDemo, and then start the operation software so that the software enters a Demo mode. BS-240 Service Manual 4-9 4 Software Introduction 4.5 Demo Software Setup 4-10 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units BS-240 Chemistry Analyzer Service Manual 5-1 5 Modules and Units 5.1 Shell Assembly 5.1 Shell Assembly 5.1.1 Main functions The shells assembly indicates the apparent structures of the whole analyzer and is designed for preventing dust and protecting the internal assemblies. It provides interfaces for each module, and is also a representation of industrial design. 5.1.2 Component Locations and FRU Details The shells assembly consists of the shielding cover, top cover, left and right panel assemblies, front cover assembly, panel, middle baffle, and rear panel. Figure 5.1 BS-240 shell assembly (1) Shielding cover hinge assembly Syringe cover assembly Auto wash tube sleeve Shielding cover assembly Middle baffle (BA24) Air spring ISE port plug for panel Sample/reagent panel Sample/reagent carousel cover assembly Wash station desk panel Air spring holder Reaction carousel panel assembly (BA24) Lamp panel Screw plug for panel Front cover assembly (BA24) Left panel assembly 5-2 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.1 Shell Assembly Figure 5.2 BS-240 shell assembly (2) Top cover Rear panel Right panel assembly Table 5.1 List of materials No BS-240 Service Manual FRU No. or Part No. Part Name Remark 1 115-036569-00 Shielding cover assembly FRU 2 042-016861-00 Middle baffle (BA24) FRU(BS-240) 3 042-016862-00 Middle baffle (BA25) FRU(BS-230) 4 M6T-010005--- Air spring FRU 5 BA40-20-72908 ISE port plug for panel / 6 043-006892-00 Sample/reagent panel FRU 7 115-036339-00 Sample/reagent carousel cover assembly FRU 8 115-036338-00 Left panel assembly FRU 9 115-036337-00 Front cover assembly (BA24) FRU(BS-240) 10 115-036334-00 Front cover assembly (BA25) FRU(BS-230) 11 BA40-20-72907 Screw plug for panel (Mould MR72907) FRU 12 043-006894-00 Lamp panel FRU 13 115-036340-00 Reaction carousel panel assembly (BA24) FRU(BS-240) 14 115-036335-00 Reaction carousel panel assembly (BA25) FRU(BS-230) 15 042-018183-00 Air spring holder / 16 043-006896-00 Wash station desk panel FRU(BS-240) 17 043-006897-00 Wash tube sleeve /(BS-240) 18 BA10-30-78156 Syringe skylight cover assembly FRU 19 BA10-20-77801 Top cover FRU 20 115-036570-00 Right panel assembly FRU 21 042-016863-00 Rear panel / 5-3 5 Modules and Units 5.1 Shell Assembly 5.1.3 Replacing sheilding cover When to do Replace the shielding cover when it is damaged. Tools Name Code Quantity Cross head screwdriver / 1 Flathead screwdriver(small) / 1 How to do 1 Cut off the power of the analyzer to ensure all the assemblies are not in the work status. 2 Open the shielding cover assembly, insert a flathead screwdriver into a gap between the circlip and the round head on the air spring head on a side of the shielding cover, and push outward the circlip to a limit of not dropping the circlip. 3 Hold the shielding cover body, and take out the air spring from the round head with a slight force, and then put the circlip back into the original position. 4 Use a cross screwdriver to remove the 4 M4×10 powder screws that fasten the hinge assembly of the shielding cover to the shielding cover assembly, so as to remove the shielding cover assembly. 5 To restore the shielding cover assembly, follow the steps mentioned above in the reversed order. Alignment and confirmation After restoring the shielding cover assembly, check the gap with the panels, and if the gap is uneven or the shielding cover interferes with the panels, adjust the position of fastening the hinges (the screw holes are rectangular holes that allow adjustment). 5.1.4 Removing and Reinstalling Left and Right Panels When to do This operation is performed when an inner component of the instrument needs repair and the left and right panels need to be removed. Tools Name Cross head screwdriver 5-4 Code / Quantity 1 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.1 Shell Assembly How to do 1 Remove the 3 screws at the back of the left panel or right panel. 2 Use one hand to hold the side panel, and use the other hand to exert a slight force of beating from the side panel arc on the upper part of the bench to the back of the instrument to push the side panel to move until the 2 stop bolts on the front end of the side panel are fully removed from the front cover bracket, and then remove the side panel. 3 To reinstall the side panel, follow the steps mentioned above in a reverse order. Alignment and confirmation The left and right panel needs to fit the top panel exactly. Otherwise, the rear side cannot fit the frame closely, which affects fastening of the screw. 5.1.5 Removing and Reinstalling Front Panel When to do This operation is performed when an inner component of the instrument needs repair and the front panel need to be removed. Tools Name Cross head screwdriver Code / Quantity 1 How to do 1 Remove the left panel and the right panel first. 2 Loosen the 2 screws (without removing) in the front of the sample/reagent panel, and then remove the lamp panel. 3 Loosen (without removing) a retaining screw of the front panel and the front beam of the frame to the maximum extent. 4 Remove the retaining screws located on both sides of the front panels and used for fastening to the frame. 5 Pull out the panel assembly towards the front of the instrument. 6 To restore the front panel assembly, follow the steps mentioned above in the reversed order. Alignment and confirmation When restoring, if the gap between the front panel and the sample/reagent panel and the lamp panel is not proper, you can remove the lamp panel, and adjust the gap by adjusting the retaining screw that fastens the front panel and the frame beam. BS-240 Service Manual 5-5 5 Modules and Units 5.1 Shell Assembly 5.1.6 Removing and Reinstalling Panels When to do This operation is performed when an inner component of the instrument needs repair and the front panel need to be removed. Tools Name Cross head screwdriver Code / Quantity 1 How to do Removing and Reinstalling BS-240 Reaction Carousel Panel 1 Remove the panel screws on the wash station panel, the reaction carousel panel, the lamp panel first, unscrew the retaining screws on the 3 parts, and remove the wash station panel and the lamp panel. 2 Loosen the knurled screws on the wash probe assembly, and remove the wash probe assembly. Handle carefully to protect the probe assembly against bending and contamination. 3 Remove the probe arm cover and the mixer arm cover. 4 Remove the reaction carousel panel assembly. Note that the panel needs to tilt at a specific angle so that it can be taken out. Handle carefully to protect the sampe probe and the mixer. 5 To restore the reaction carousel panel assembly, follow the steps mentioned above in the reversed order. Removing/Reinstalling Sample/Reagent disk panel 1 Remove the panel screws on the reaction carousel panel and the lamp panel, and unscrew the retaining screw. 2 Remove the sample/reagent carousel cover assembly from the sample/reagent carousel panel. 3 Remove the sample/reagent panel. 4 To restore the reaction carousel panel assembly, follow the steps mentioned above in the reversed order. Alignment and confirmation To restore the reaction carousel panel, adjust the gap evenly before tightening the screw. 5-6 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.1 Shell Assembly 5.1.7 Replacing Air Spring When to do Replace the air spring if it loses elasticity or fails. Tools Name Flathead screwdriver(small) Code / Quantity 1 How to do 1 Open the shielding cover, insert a flathead screwdriver into a gap between the circlip and the round head on the air spring head, and push outward the circlip to a limit of not dropping the circlip. 2 Grasp the air spring cylinder with a hand, and pull the air spring out. The undamaged air spring head is still usable. 3 Install a new air spring (only the air spring cylinder needs to be replaced). Alignment and confirmation The cylinder end of the air spring needs to located at the side of the shielding cover. After the air spring is replaced, open and close the shielding cover repeatedly for several times so that the air spring cylinder is sufficiently lubricated. 5.1.8 Cleaning Fans When to do When much dust is accumulated on the fans, or it has been 1 year since the last maintenance. Maintenance Tools Name Code Quantity Cross head screwdriver / 1 Suction cleaner or Hair brush / 1 Maintenance Procedure 1 Place the analyzing unit power to the OFF position. 2 If the reagent refrigeration fan and PCB fan need cleaning, remove the left panel assembly. 3 If the system radiating fan and the power supply radiating fan need cleaning, remove the rear panel. 4 If the lamp fan needs cleaning, remove the right panel assembly. BS-240 Service Manual 5-7 5 Modules and Units 5.1 Shell Assembly 5 Manually rotate the fan and remove the dust on it. 6 Install the removed panels. 7 Turn on the analyzing unit. Alignment and confirmation Check if the fans work normally. 5-8 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.2 Frame Assembly 5.2 Frame Assembly 5.2.1 Main functions The frame assembly is a basic unit for installing and supporting the components of the analyzer. Figure 5.3 Locations of components on the instrument Sample Probe Movement Assembly Mixer Movement Assembly Shielding Cover Warning Sample/reagent carousel cover Sample/reagent panel Reagent bar code assembly Wash station desk panel Reaction carousel panel Lamp panel Frame Assembly Reaction Carousel Assembly Left panel assembly BS-240 Service Manual Sample/reagent carousel assembly Front cover assembly Reagent refrigeration board and the bracket 5-9 5 Modules and Units 5.2 Frame Assembly 5.2.2 Component Locations and FRU Details Figure 5.4 Frame assembly Dust screen retaining screw BA24 rear dust screen BA24 front dust screen Dust screen retaining screw Table 5.2 List of materials No FRU No. or Part No. Part Name Remark 1 048-006111-00 BA24 front dust screen FRU 2 048-006112-00 BA24 rear dust screen FRU 5.2.3 Replacing Front and Rear Dust Screens When to do The front and rear dust screens need to be replaced when they need cleaning or are damaged. Tools Name Cross head screwdriver 5-10 Code / Quantity 1 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.2 Frame Assembly Exploded view for installation See the schematic diagram of the frame assembly. How to do 1 Switch off the power supply of the whole unit. 2 Remove the right panel assembly. 3 Use a cross head screwdriver to loosen the dust screen retaining screw M3×10 and cross recessed pan head screw. 4 Remove the dust screens. 5 Replace or clean the dust screens. 6 Restore the assembly in the reversed order. Alignment and confirmation Make sure that the dust screen completely covers the air inlet. BS-240 Service Manual 5-11 5 Modules and Units 5.3 Reaction Carousel Assembly 5.3 Reaction Carousel Assembly 5.3.1 Main functions Reaction carousel assembly is situated at the right front side of the analyzer, including the body of the reaction carousel, reaction carousel drive assembly, reaction carousel chamber, reaction carousel sensor, and the reaction carousel motor assembly. It is used to complete various actions, such as carrying and rotating the reaction cuvette to the specified position, adding sample/reagent to the cuvette, mixing and washing, together with the sample probe, the mixer and the 4-phase wash probes. It, as well as, provides a constant temperature for the reaction cuvette for analysis. Figure 5.5 Locations of assemblies on the instrument Sample Probe Movement Assembly Shielding Cover Warning Sample/reagent carousel cover Mixer Movement Assembly Wash station desk panel Sample/reagent panel Reagent bar code assembly Reaction carousel panel Lamp panel Frame Assembly Reaction Carousel Assembly Left panel assembly 5-12 Sample/reagent carousel assembly Front cover assembly Reagent refrigeration board and the bracket BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.3 Reaction Carousel Assembly 5.3.2 Locations and FRU details The reaction carousel assembly is located in the right front side of the instrument. Figure 5.6 Reaction carousel assembly Upper cover assembly Optical assembly Reaction carousel motor assembly BS-240 Service Manual Sensor assembly Reaction carousel heat chamber assembly 5-13 5 Modules and Units 5.3 Reaction Carousel Assembly Temperature protection switch wire (for reaction carousel) Heater cable above reaction carousel Cuvettes Cover Reaction Carousel Body Assembly M2.5×8 cross countersunk head screw M3× cross pan head combination screw Temperature control fan cable M5×20 hexagon screw M3×14 cross recessed pan head screw Reaction carousel temperature sensor cable Reaction chamber Stand bar of fan Heater cable under reaction carousel Thermal Guard 60C 2Lead15.6mm M3×6 cross recessed pan head screw Heater baffle plate M4×8 cross pan head screw Reaction carousel motor assembly Reaction carousel support Reaction Carousel Sensor Assembly Table 5.3 List of materials No Part Name Remark 1 009-006306-00 Reaction carousel temperature sensor cable FRU 2 115-006048-00 Motor assembly FRU 3 BA10-20-78144 Temperature control fan cable FRU BA10-21-78143 Heater cable under reaction carousel FRU BA10-21-78142 Heater cable above reaction carousel FRU BA10-21-78147 Temperature protection switch wire (for reaction carousel) FRU 7 009-002204-00 Correlative photocoupler (S) FRU 8 M6C-020007--- Synchronous TBN180XL037 FRU 4 5 6 5-14 FRU No. or Part No. cog belt, BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.3 Reaction Carousel Assembly 5.3.3 Replacing Cables of Upper Cover Assembly When to do The temperature in the heat chamber of the reaction carousel is abnormal, and it is checked that the heater cable on the reaction carousel or the temperature protection switch cable of the reaction carousel is faulty. In this case, the faulty part needs to be replaced. Tools Name Cross head screwdriver Code / Quantity 1 Exploded view for installation Figure 5.7 Upper cover assembly cables M4×12 cross pan head screw M4×8 cross countersunk head screw Reaction carousel anchor plate Temperature protection switch wire (for reaction carousel) Heater cable above reaction carousel Cover How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, loosen the knurled screws on the wash probe assembly, remove the wash probe assembly, and place the probe properly against bending. Remote the sample probe to the reagent aspirating port position, remote the mixer to the wash well position, unscrew the M4×8 cross recessed pan head padded screw, remove the BA24 reaction carousel panel, lamp panel, wash panel, unscrew the 3 M4×8 cross recessed pan head screws for fastening the right panel assembly, and remove the right panel assembly. BS-240 Service Manual 5-15 5 Modules and Units 5.3 Reaction Carousel Assembly 3 Use a cross head screwdriver to unscrew the 4 M4×12 cross recessed combination screws that fasten the upper cover assembly, unplug the heater connection cable on the reaction carousel and the temperature protection switch cable of the reaction carousel, and remove the upper cover assembly. 4 Unscrew the 4 M4×8 cross pan head combination screws that fasten the anchor plate of the reaction carousel, and remove the anchor plate of the reaction carousel. 5 Remove the faulty heater cable on the reaction carousel or the faulty temperature protection switch cable of the reaction carousel. 6 To replace the heater cable on the reaction carousel, apply thermal paste evenly on the heater cable of the reaction carousel, and apply a small amount of hot melt adhesive at the outlet of the heater cable on the reaction carousel to fasten the leadout cable. 7 To replace the temperature protection switch cable of the reaction carousel, apply thermal paste evenly in the installation slot on the panel, and apply a small amount of hot melt adhesive at the outlet of the temperature protection switch cable of the reaction carousel to fasten the lead-out cable. 8 Restore the components in the reversed order. Alignment and confirmation After replacing the heater cable on the reaction carousel or the temperature protection switch cable of the reaction carousel, configure the sensor-related parameters on software, and perform the heat insulation performance test. For sensor parameter configuration, see 8.10.1; for reaction carousel temperature test, see 8.10.2; for reaction carousel temperature curve, see 8.10.2. 5.3.4 Replacing Heat Chamber Assembly Cables When to do The temperature in the heat chamber of the reaction carousel is abnormal, and it is checked that the heater cable under the reaction carousel, or the Thermal Guard 60C 2Lead15.6mm, or the temperature sensor cable, or the temperature control chamber fan cable is faulty. In this case, the faulty part needs to be replaced. Tools Name 5-16 Code Quantity Cross head screwdriver / 1 Hexagon wrench / 1 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.3 Reaction Carousel Assembly Exploded view for installation Figure 5.8 Assembling/disassembling heat chamber assembly cable of reaction carousel Upper cover assembly M2.5×8 cross countersunk head screw M3×6 cross pan head combination screw M3×14 cross recessed pan head screw Temperature control fan cable Stand bar of fan Reaction carousel temperature sensor cable M5×20 hexagon screw Reaction chamber Heater cable under reaction carousel Thermal Guard 60C 2Lead15.6mm M3×6 cross recessed pan head screw Heater baffle plate M4×8 cross pan head screw Reaction carousel support How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, loosen the knurled screws on the wash probe assembly, remove the wash probe assembly, and place the probe properly against bending. Remote the sample probe to the reagent aspirating port position, remote the mixer to the wash well position, unscrew the M4×8 cross recessed pan head padded screw, remove the BA24 reaction carousel panel, lamp panel, wash panel, unscrew the 3 M4×8 cross recessed pan head screws for fastening the right panel assembly, and remove the right panel assembly. 3 Use a cross head screwdriver to unscrew the 4 M4×12 cross recessed combination screws that fasten the upper cover assembly, unplug the heater connection cable on the reaction carousel and the temperature protection switch cable of the reaction carousel, and remove the upper cover assembly and place it properly BS-240 Service Manual 5-17 5 Modules and Units 5.3 Reaction Carousel Assembly 4 Disassemble the BS-240 reaction carousel assembly: Loose the anchor plate of the reaction cuvette, remove the reaction cuvette segment, and place them properly. Loosen the 3 M3×12 hexagon screws that fasten the reaction carousel assembly, and remove the reaction carousel assembly, as shown in the following diagram: Figure 5.9 Assembling/disassembling reaction carousel assembly Reaction cuvette Reaction cuvette segment anchor plate M3×12 hexagon screw 5 Fan and bracket Reaction carousel temperature sensor cable Reaction Carousel Body Assembly To replace the temperature sensor cable of the reaction carousel, use a cross head screwdriver to loosen the M3×8 cross countersunk head screw, unplug the temperature sensor cable of the reaction carousel, remove the temperature sensor cable of the reaction carousel directly, and replace it with a new temperature sensor cable of the reaction carousel and plug it properly. 6 To replace the heat chamber fan cable, unplug the heat chamber fan cable, use a cross head screwdriver to unscrew 2 M3×6 cross recessed pan head combination screws, remove the stand bar of the fan together with the fan, and install a new fan onto the stand bar and plug it properly. (Note: When installing the fan, please make sure that the label faces the stand bar, and the cable of the fan is led out from the lower right side) 7 To change the heater cable under the reaction carousel or the Thermal Guard 60C 2Lead15.6mm, remove 1 M4×12 screw and two M4×16 hexagon screws that fasten the optical assembly, unplug the cable that connects the motor and the sensor, loosen the 3 M5X20 hexagon screws that fasten the reaction chamber, unplug the related cables, and remove the reaction chamber. 8 To replace the Thermal Guard 60C 2Lead15.6mm, put the reaction chamber upside down, and loosen the 2 M3×6 cross recessed pan head screws directly to perform replacement. 5-18 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.3 Reaction Carousel Assembly 9 To replace the heater cable under the reaction carousel, loosen 10 M4×8 cross recessed pan head combination screws that fasten the baffle plate of the heater, remove the baffle plate of the heater, and replace the heater cable under the reaction carousel. Apply thermal paste evenly on the heater cable under the heater of the reaction carousel, and apply a small amount of hot melt adhesive at the outlet of the heater cable under the reaction carousel to fasten the lead-out cable. 10 Restore the assembly in reserved order. Try to align the heat chamber with the rotor of the reaction carousel when installing the heat chamber. Alignment and confirmation After replacing the heater cable under the reaction carousel, the Thermal Guard 60C 2Lead15.6mm, and the temperature sensor cable of the reaction carousel, reconfigure the related sensor parameters on the software and perform the heat insulation performance test. For sensor parameter configuration, see 8.10.1; for reaction carousel temperature test, see 8.10.2; for reaction carousel temperature curve, see 8.10.2. 5.3.5 Replacing Cables of Home Position Sensor and Coder Sensor When to do Replace the home position sensor or the coder sensor of the reaction carousel when they fail. Tools Name Code Quantity Cross head screwdriver / 1 Hexagon wrench / 1 Exploded view for installation Figure 5.10 Replacing cables of home position sensor and coder sensor M3×6 cross pan head screw with washer Coder sensor Home position sensor Mounting plate of sensor Sensor bracket BS-240 Service Manual M3×8 hexagon socket head cap screw Install sensor in left screw hole of bracket M3×6 cross recessed pan head screw 5-19 5 Modules and Units 5.3 Reaction Carousel Assembly NOTE: Install the sensors according to the identifications on the cable connectors and the instrument connectors: "RCD--PHO-C" for coder sensor and "RCD--PHO-O" for home position sensor. Do not tighten the sensor screws with excessive torque force, in order to avoid damaging the sensors. Make sure to install the home position sensor in the correct position. How to do Disassembling and assembling the coder sensor. 1 Switch off the power of the analyzer. 2 Unscrew 3 M4×8 cross recessed pan head screws that fasten the right panel assembly, remove the right panel assembly, and then unscrew 2 M4×8 cross recessed pan head combination screws that fasten the AD bracket assembly, and put the AD bracket assembly aside (without removing the cable of the AD box). 3 Unplug the cable of the coder sensor. 4 Loosen the 2 M3×6 cross combination screws that fasten the mounting plate of the sensor, and remove the coder sensor cable together with the mounting plate. 5 Loosen the 2 cross pan head screws on the coder sensor, and remove the coder sensor from the mounting plate. 6 Fix the new coder sensor on the mounting plate using two M3×6 cross pan head screws. 7 Use 2 M3×6 cross combination screws to attach the mounting plate of the sensor to the bracket of the sensor, adjust the height of the mounting plate, make sure that the coder lie in the middle of the coder sensor, and then tighten the screws of the mounting plate. 8 Plug the cable of the coder sensor. 9 After aligning and checking the sensor assembly of the reaction carousel, restore the assembly in the reversed order of steps 1~2. Disassembling and assembling home position sensor 1 Switch off the power of the analyzer. 2 Unscrew 3 M4×8 cross recessed pan head screws that fasten the right panel assembly, remove the right panel assembly, and then unscrew 2 M4×8 cross recessed pan head combination screws that fasten the AD bracket assembly, and put the AD bracket assembly aside (without removing the cable of the AD box). 3 5-20 Unplug the connectors of the reaction carousel home position sensor and coder sensor. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.3 Reaction Carousel Assembly 4 Unscrew the three M3X8 hexagon socket head cap screws with spring and flat washers on the reaction carousel sensor assembly to remove it. Do not drop the screws into the analyzer. 6 Unscrew 2 M3×6 cross pan head screws on the cables of carousel home sensor to remove it from the bracket of the sensor. 7 Fix the new cables of carousel home sensor onto the sensor bracket using 2 M3×6 cross pan head screws. 8 Fix the reaction carousel sensor assembly on the large bottom plate using 3 M3×8 hexagon socket head cap screws with spring and flat washers. Rotate the reaction carosuel with a hand, and prevent the zero position baffle plate of the coder from interfering with the sensor. 9 Connect the home position sensor and the coder sensor. 10 After aligning and checking the sensor assembly of the reaction carousel, restore the assembly in the reversed order of steps 1~2. Alignment and confirmation Align the photoelectric signal collecting position according to 8.3.1, and align the circumferential position of the reaction carousel assembly, the circumferential position of the sample/reagent carousel assembly, the horizontal position of the sample probe and the mixer, and the horizontal position of the wash station according to 8.4.1~8.4.4. 5.3.6 Replacing Motor Assembly and Synchronous Belt When to do Replace the reaction carousel motor or synchronous belt when they are damaged. Tools Name BS-240 Service Manual Code Quantity Cross head screwdriver / 1 Hexagon wrench / 1 5-21 5 Modules and Units 5.3 Reaction Carousel Assembly Exploded view for installation Figure 5.11 Reaction carousel motor assembly M4×10 hexagon socket head cap screw + spring washer Motor Motor shock pad M5×12 hexagon socket head cap screw + spring/flat washer WARNING During removing and installation, avoid dropping screws and washers into the analyzer. How to do Replace the motor 1 Switch off the power of the analyzer. 2 Unscrew 3 M4×8 cross recessed pan head screws that fasten the left panel assembly and those that fasten the right panel assembly, remove the left and right panel assemblies, unscrew M4×8 cross countersun head screws that fasten the front housing of the BS-240, and remove the front housing of the BS-240. 3 Disconnect the motor cable, unscrew the 4 M5×12 hexagon socket head screws with spring washer on the reaction carousel motor assembly, remove the motor assembly and avoid dropping screws into the analyzer. 4 Fix a new motor assembly to the large bottom plate using 4 M5×12 hexagon socket head cap screws with spring washer. Tighten the belt and screws and connect the cable of the motor. 5 Restore the components in the reversed order. Replacing the synchronous belt 1 5-22 Switch off the power of the analyzer. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.3 Reaction Carousel Assembly 2 Open the shielding cover, loosen the knurled screws on the wash probe assembly, remove the wash probe assembly, and place the probe properly against bending. Remote the sample probe to the reagent aspirating port position, remote the mixer to the wash well position, unscrew the M4×8 cross recessed pan head padded screw, remove the BA24 reaction carousel panel, lamp panel, wash panel, unscrew the 3 M4×8 cross recessed pan head screws for fastening the left panel assembly and those for fastening the right panel assembly, and remove the left and right panel assemblies. Unscrew the M4×8 cross countersunk head screw that fastens the front housing of the BS-240, and remove the front housing of the BS-240. 3 Use a cross head screwdriver to unscrew the 4 M4×12 cross recessed combination screws that fasten the upper cover assembly, unplug the heater connection cable on the reaction carousel and the temperature protection switch cable of the reaction carousel, and remove the upper cover assembly and place it properly 4 Loose the anchor plate of the reaction cuvette, remove the reaction cuvette segment, and place them properly. Loosen the 3 M3×12 hexagon screws that fasten the reaction carousel assembly, and remove the reaction carousel assembly. 5 Remove 1 M4×12 screw and two M4×16 hexagon screws that fasten the optical assembly, unplug the cable that connects the motor and the sensor, loosen the 3 M5X20 hexagon screws that fasten the reaction chamber, unplug the related cables, and remove the reaction chamber. 6 Disconnect the motor cable, unscrew the 4 M5×12 hexagon socket head screws with spring washer on the reaction carousel motor assembly, remove the motor assembly and avoid dropping screws into the analyzer. 7 Install a new synchronous belt. 8 Fix a reaction carousel motor assembly to the large bottom plate using 4 M5×12 hexagon socket head cap screws with spring washer. Tighten the belt and screws and connect the cable of the motor. 9 Restore the assembly in reserved order. Try to align the heat chamber with the rotor of the reaction carousel when installing the heat chamber. Alignment and confirmation N/A 5.3.7 Replacing Reaction Cuvette See 5.9.8 for the methods and steps of replacing reaction cuvette. BS-240 Service Manual 5-23 5 Modules and Units 5.4 Sample/reagent carousel assembly 5.4 Sample/reagent carousel assembly 5.4.1 Main functions The sample/reagent carousel assembly is situated at the left front of the analyzer, and includes the sample/reagent carousel body assembly, sample/reagent refrigeration assembly, sample/reagent carousel drive assembly, motor assembly, coder photocoupler assembly, and bar code reader (optional). It is used to complete various actions, for example, carry and rotate the reagent bottle with reagent and the sample container with samples, so that the reagent/sample reach the specified position, add samples together with sampling assembly. It provides a refrigerating environment, and the reagents stored in such environment can be kept stable with little volatilization. For details, please refer to the following contents. 1 Holding reagent/sample: Holding specified number of reagent bottles/sample containers of certain specification, and cooperating with the dispenser to aspirate reagent/sample from the bottles and dispense it into reaction cuvettes. 5-24 2 Refrigerating reagent: Provide 24-hour refrigeration and constant temperature environment (2-10°C) to keep the reagent stable with little volatilization.. 3 Sequential feeding: The drive assembly drives the reagent carousel to carry certain reagent bottle in the given order to the aspirate position for aspiration. 4 Input reagent/sample data automatically: Bar code reader (optional). BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly Figure 5.12 Locations of assemblies on the instrument Sample Probe Movement Assembly Mixer Movement Assembly Shielding Cover Sample/reagent carousel cover Wash station desk panel Sample/reagent panel Reagent bar code assembly Reaction carousel panel Lamp panel Frame Assembly Reaction Carousel Assembly Left panel assembly Sample/reagent carousel assembly Front cover assembly Reagent refrigeration board and the bracket 5.4.2 Assembly Locations and FRU Details The sample/reagent carousel assembly is located in the front left of the instrument. Figure 5.13 Sample/reagent carousel assembly BS-240 Service Manual 5-25 5 Modules and Units 5.4 Sample/reagent carousel assembly 1 11 10 2 9 3 4 5 8 6 7 Table 5.4 List of materials No 5-26 FRU No. or Part No. Part Name Remark 1 115-036341-00 Sample/reagent carousel body assembly FRU 2 024-000110-00 Sensor temperature 5Kohm B3470K with threads FRU 3 082-000855-00 O-shaped 48.7X2.65 silicon rubber black A70 / 4 082-000856-00 O-shaped 65X2 silicon rubber black A70 / 5 BA40-21-61655 Semi-conductive Peltier cable FRU 6 009-002204-00 Correlative photocoupler (S) FRU 7 115-036317-00 Sample/reagent carousel motor assembly FRU 8 M6C-020007--- Synchronous cog belt, TBN180XL037 FRU 9 BA10-20-78146 Cables of reagent refrigeration and radiating fan FRU 10 115-036320-00 Reagent assembly heater FRU 11 023-000738-00 Built-in bar code reader MS-3 High density laser D-Sub FRU 12 BA20-20-75214 Cuvette holder base 1 FRU 13 115-036342-00 Handle assembly FRU 14 BA30-20-15100 Rubber pad 1 FRU 15 BA30-20-15210 Transparent shield rubber pad FRU 16 041-022590-00 Sample tube adapter FRU carousel anti-fogging BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly 5.4.3 Replacing Sample/Reagent Carousel Body Assembly When to do Replace the sample/reagent carousel body assembly when it is damaged. Tools Name Code Quantity Cross head screwdriver / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.14 Reagent carousel body assembly Sample/reagent carousel body assembly How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, rotate the sample probe to the reaction carousel side, and remove the sample/reagent carousel cover assembly. 3 Lift the handlebar to remove the sample/reagent carousel body assembly. 4 Align the locating holes on the new sample carousel body with the stop bolt, and pinch the handlebar flat. 5 Restore the sample/reagent carousel cover, and close the shielding cover. Alignment and confirmation N/A BS-240 Service Manual 5-27 5 Modules and Units 5.4 Sample/reagent carousel assembly 5.4.4 Replacing Temperature Sensor When to do Replace the temperature sensor when it fails. Tools Name Code Quantity Cross head screwdriver / 1 Flathead screwdriver / 1 Hexagon wrench / 1 Glue gun / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.15 Disassembling/assembling temperature sensor Sample/reagent carousel body assembly Refrigeration sleeve M3×10 cross countersunk head screw M4×10 hexagon socket cap head screw Rubber pad 1 Abrasion-resistant pad Reagent refrigeration chamber Rubber cushion Temperature sensor How to do 1 5-28 Switch off the power of the analyzer. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly 2 Open the shielding cover, remove the sample/reagent carosuel cover, and remove the left panel assembly, front cover assembly, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the sample probe from colliding during the deatching. 3 Lift the handlebar to remove the sample/reagent carousel body assembly. 4 Remove the reagent refrigeration board and bracket, disconnect the drain tube of the reagent refrigeration compartment from the reagent refrigeration compartment, remove the temperature sensor connector, and remove the reagent bar code assembly if the bar code is configured. 5 Use a cross head screwdriver to remove 6 M3×10 cross countersunk head screws that fasten the refrigeration sleeve, and remove the refrigeration sleeve and the rubber pad 1. 6 Use the hexagon wrench to remove the 8 M4×10 hexagon socket head cap screws that fasten the reagent refrigeration compartment, remove the abrasion-resistant pad, and remove the reagent refrigeration compartment. 7 Clear the sealing glue in the installation position of the temperature sensor, and remove the temperature sensor and the rubber pad. 8 Sheathe the rubber pad into a new temperature sensor, load them into the reagent refrigeration compartment, and seal the installation hole with the sealing glue. 9 Apply thermal paste on the installation surface of the reagent refrigeration compartment and the refrigeration aluminum block, apply the sealing glue on the abrasion-resistant pad, and use 8 M4×10 hexagon socket head cap screws to install the reagent refrigeration compartment onto the refrigeration aluminum block. Try to align the reagent refrigeration compartment with the rotor. 10 Sheathe the rubber pad 1 into the refrigeration sleeve, apply the sealing glue to the fitting surface of the reagent refrigeration compartment, and use 6 M3×6 cross countersunk head screws to fix the refrigeration sleeve. 11 Install the disassembled parts in reversed order of steps 2~4. NOTE When filling the installation hole of the temperature sensor with the sealing glue, completely fill the cable hole with glue. Wipe off overflowing glue when filling the abrasion-resistant pad. Alignment and confirmation 1 Re-align the sample probe to the positions of the outer ring sample tube and the middle and inner ring reagent bottle (BA60-J07 pseudo probe, BA48-J10 sample position alignment fixture, and BA10-J16 reagent center align with the alignment fixture). BS-240 Service Manual 5-29 5 Modules and Units 5.4 Sample/reagent carousel assembly 2 Re-align the bar code scan position (light adjustment tool of BA24-J03 BA24 bar code scan). 5.4.5 Replacing Peltier and Sealing Ring When to do Replace the Peltier when it fails. Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Glue gun / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.16 Disassembling and assembling Peltier and sealing ring M3×10 cross countersunk head screw Refrigeration sleeve Rubber pad 1 Sample/reagent carousel body assembly M4×10 hexagon socket cap head screw Abrasion-resistant pad Reagent refrigeration chamber Refrigeration aluminum block M4×8 cross countersunk head screw Refrigeration aluminum block O-shaped 48.7×2.65 sealing ring Refrigeration fixing base O-shaped 65×2 sealing ring Peltier 5-30 M4×8 cross countersunk head screw O-shaped 48.7×2.65 sealing ring M5×20 hexagon socket head cap screws and spring/flat washer Refrigeration fixing base O-shaped 65×2 sealing ring Peltier BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carosuel cover, and remove the left panel assembly, front cover assembly, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the sample probe from colliding during the deatching. 3 Lift the handlebar to remove the sample/reagent carousel body assembly. 4 Remove the reagent refrigeration board and bracket, disconnect the drain tube of the reagent refrigeration compartment from the reagent refrigeration compartment, detach the temperature sensor connector, and detach the reagent bar code assembly if the bar code is configured. 5 Use a cross screwdriver to remove 6 M3×10 cross countersunk head screws that fasten the refrigeration sleeve, and remove the refrigeration sleeve and the rubber pad 1. 6 Use the hexagon wrench to remove the 8 M4×10 hexagon socket head cap screws that fasten the reagent refrigeration compartment, remove the abrasion-resistant pad, and remove the reagent refrigeration compartment. 7 Remove the 2 M5×20 hexagon socket head cap screws, the elastomer and the flat washer that are used to fasten the refrigeration base on the Peltier to be replaced, and remove the refrigeration base that carry a refrigeration aluminum block, so that the Peltier and the O-shaped 65×2 sealing ring can be replaced. 8 To replace the O-shaped 48.7×2.65 sealing ring, remove the 4 M4×8 cross countersunk head screws that fasten the refrigeration aluminum block, and replace the O-shaped 48.7×2.65 sealing ring. 9 Apply thermal pate on both sides of the substitute Peltier, install the Peltier according to the direction before disassembling, and plug the cable. Install the substitute sealing ring. 10 Use 4 M4×8 cross countersunk head screws to fix the refrigeration aluminum block onto the refrigeration base, and then use 2 M5×20 hexagon socket head cap screws, the elastomer and the flat washer to fix the refrigeration base onto the Peltier. The direction needs to be consistent with the direction before disassembling. 11 Apply thermal paste on the installation surface of the reagent refrigeration compartment and the refrigeration aluminum block, apply the sealing glue on the abrasion-resistant pad, and use 8 M4×10 hexagon socket head cap screws to install the reagent refrigeration compartment onto the refrigeration aluminum block. Try to align the reagent refrigeration compartment with the rotor. 12 Sheathe the rubber pad 1 into the refrigeration sleeve, apply the sealing glue to the fitting surface of the reagent refrigeration compartment, and use 6 M3×6 cross countersunk head screws to fix the refrigeration sleeve. 13 Install the disassembled parts in reversed order of steps 2~4. BS-240 Service Manual 5-31 5 Modules and Units 5.4 Sample/reagent carousel assembly NOTE When filling the installation hole of the temperature sensor with the sealing glue, completely fill the cable hole with glue. Wipe off overflowing glue when filling the abrasion-resistant pad. Alignment and confirmation 1 Re-align the sample probe to the positions of the outer ring sample tube and the middle and inner ring reagent bottle (BA60-J07 pseudo probe, BA48-J10 sample position alignment fixture, and BA10-J16 reagent center align with the alignment fixture). 2 Re-align the bar code scan position (adjustment tool of the bar code scan light of BA24-J03 BA24). 5.4.6 Replacing Cables of Home Position Sensor and Coder Sensor When to do Replace the home position sensor or the coder sensor of the sample/reagent carousel when they fail. Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.17 Disassembling/assembling correlative photocoupler (S) cables M3×6 cross pan head screw with washer M3×6 cross recessed pan head screw Mounting plate of sensor Coder photocoupler M3×6 hexagon socket cap head screw M3×6 cross recessed pan head screw M3×6 hexagon socket head cap screws and spring/flat washer Photocoupler bracket Home position photocoupler How to do 1 5-32 Switch off the power of the analyzer. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly 2 Remove the left panel assembly and the front panel assembly. 3 Remove the reagent refrigeration board and bracket, and unplug home position photocoupler of the reagent carousel and the coder photocoupler. 4 Unscrew the 3 M3×6 hexagon socket head cap screws and the spring/flat washer that are used to fasten the coder photocoupler assembly, and remove the coder photocoupler assembly. 5 Loosen the 2 M3×6 cross pan head combination screws on the photocoupler mounting plate, and then remove the photocoupler mounting plate together with the coder photocoupler. 6 Loosen the 2 M3×6 cross pan head screws that fasten the coder photocoupler, and remove the coder photocoupler from the sensor mounting plate. 7 Use 2 M3×6 cross pan head screws to fix the new coder photocoupler onto the sensor mounting plate. 8 Use 2 M3×6 cross recessed pan head combination screws to fix the sensor with the coder photocoupler onto the photocoupler bracket, without tightening them. 9 Unscrew the 2 M3×6 cross recessed pan head screws of the photocoupler in the home position of the reagent carousel, and remove the photocoupler from photocoupler bracket. 10 Use 2 M3×6 cross recessed pan head screws to fix a new photocoupler cable in the home position of the reagent carousel onto the photocoupler bracket 11 Use 3 M3×6 hexagon socket head cap screws to fix the coder photocoupler assembly onto a large bottom plate, push the photocoupler bracket to a stop position of the two hexagon screws, fit the bracket closely, and tighten the retaining screws. Adjust the sensor mounting plate to make the coder lie in the middle of the coder photocupler, and then tighten the screws on the mounting plate. 12 Plug the connectors of the photocoupler in the home position of the reagent carousel and the coder photocoupler. 13 After aligning the position of the coder photocoupler assembly, restore the assembly in the reversed order. Note: To replace only the coder photocoupler, omit steps 9 and 10. To replace only the photocoupler in the home position of the reagent coupler, omit steps 5 to 8. Alignment and confirmation After replacing the sensor and the coder sensor in the home position on the sample/reagent carousel, refer to sections 8.4~8.5 to align the circumferential position of the sample/reagent carousel assembly and align the horizontal position of the sample probe on the sample/reagent carousel, and, if bar code is configured, align the bar code scan position. BS-240 Service Manual 5-33 5 Modules and Units 5.4 Sample/reagent carousel assembly 5.4.7 Replacing Sample/Reagent Carousel Motor Assembly When to do Replace the motor when it fails. Tools Name Code Quantity Hexagon wrench / 1 Cross screwdriver / 1 M3×20 hexagon socket head cap screws and spring/flat washer / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.18 Disassembling and assembling sample/reagent carousel motor assembly Synchronous belt M4×10 hexagon socket head cap screws and spring/flat washer M3×20 hexagon socket cap head screw hole Sample/reagent carousel motor assembly How to do 1 Switch off the power of the analyzer. 2 Remove the left panel assembly and the front panel assembly. 3 Unplug the motor cable, use a hand to check the tension of the synchronous belt and keep a record. 4 Use the hexagon wrench to remove the 4 M4×10 hexagon socket head cap screws and spring/flat washer that are used to fasten the sample/reagent carousel motor assembly, and remove the assembly from the synchronous belt. 5-34 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly 5 Replace the assembly with a new sample/reagent carousel motor assembly, sheathe the pulley of the assembly into the synchronous belt, and then tighten the M4×10 hexagon socket head cap screws and spring/flat washer. Pay attention to the motor direction. 6 Pass 1 M3×20 hexagon socket head cap screw with a washer (either spring or flat washer, depending on requirements) through a tension hole, fasten the screw into the motor mounting plate, strain the belt to the original status, and tighten the motor assembly screw. Remove the M3×20 hexagon socket screw, and plug the motor cable. 7 Assemble and reset the front panel assembly and the left panel assembly. NOTE Make sure that the direction of the motor is correct. Alignment and confirmation N/A 5.4.8 Replacing the synchronous belt When to do Replace the belt when it is damaged. Tools Name BS-240 Service Manual Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Glue gun / 1 Rubber gloves for medical use / 1 pair M3×20 hexagon socket head cap screws and spring/flat washer / 1 5-35 5 Modules and Units 5.4 Sample/reagent carousel assembly Exploded view for installation Figure 5.19 Disassembling/assembling synchronous belt Sample/reagent carousel body assembly M3×10 cross countersunk head screw Refrigeration sleeve Rubber pad 1 Rotor Gap Reagent refrigeration chamber Synchronous belt M3×20 hexagon screw hole Sample/reagent carousel motor assembly Synchronous belt replacing space M4×10 hexagon socket cap head screw How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carosuel cover, and remove the left panel assembly, front cover assembly, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the sample probe from colliding during the deatching. 3 Lift the handlebar to remove the sample/reagent carousel body assembly. 4 Remove the reagent refrigeration board and the bracket. 5 Use a cross screwdriver to remove 6 M3×10 cross countersunk head screws that fasten the refrigeration sleeve, and remove the refrigeration sleeve and the rubber pad 1. 6 Draw an aligned mark line on the motor mounting plate and the large bottom plate (in the tension direction of the motor assembly to facilitate use after the synchronous belt is replaced), loosen the 4 M4×10 hexagon socket head cap screws that fasten the motor assembly of the sample/reagent carousel until the motor assembly can move, without taking out the screws. 7 Remove the synchronous belt. If the belt is broken or can be cut off, remove the belt. If the belt is not broken and needs to keep the current status, remove the belt from the gap between the reagent refrigeration compartment and the rotor. 5-36 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly 8 Sheathe a new synchronous belt into the rotor, and feed the belt slowly into the gap between the reagent refrigeration compartment and the rotor. At the belt replacing operation space and at the motor end, sheathe the belt into the motor pulley and rotor pulley. 9 Pass 1 M3×20 hexagon socket head cap screw with a washer (either spring or flat washer, depending on requirements) through a tension hole, fasten the screw into the motor mounting plate, strain the belt until it is aligned with the mark line drawn before the motor mounting plate is loosened, and tighten the motor assembly screw. Remove the M3×20 hexagon screw. 10 Sheathe the rubber pad 1 into the refrigeration sleeve, apply the sealing glue to the fitting surface of the reagent refrigeration compartment, and use 6 M3×6 cross countersunk head screws to fix the refrigeration sleeve. 11 Install the disassembled parts in reversed order of steps 2~4. NOTE Pass the synchronous belt through the gap between the reagent refrigeration compartment and the rotor slowly, and avoid applying a large force. Alignment and confirmation N/A 5.4.9 Replacing Reagent Refrigeration and Radiating Fans When to do Replace the fan when it fails. Tools Name BS-240 Service Manual Code Quantity Hexagon wrench / 1 Rubber gloves for medical use / 1 pair Cross screwdriver / 1 5-37 5 Modules and Units 5.4 Sample/reagent carousel assembly Exploded view for installation Figure 5.20 Disassembling/assembling reagent refrigeration and radiating fans M4×45 cross recessed pan head screw and flat washer φ92 fan screen cover Reagent refrigeration/radiating fan How to do 1 Switch off the power of the analyzer. 2 Remove the left panel assembly and the front panel assembly. 3 Remove the connector of the fan that needs to be replaced. 4 Remove the 4 M4×45 cross recessed pan head screws and the flat washer of the fan that needs to be replaced, and remove the fan screen and the fan. 5 Install a new fan. The installation direction of the fan needs to be consistent with the original direction, and tidy the cable and the connector. 6 Install the front panel assembly and the left panel assembly. Alignment and confirmation After powering on the instrument, check that the cooling fan is blowing wind outwards, 5.4.10 Replacing Bar Code Reader When to do Replacing the bar code reader when it fails. Tools Name 5-38 Code Quantity Hexagon wrench / 1 Rubber gloves for medical use / 1 pair Cross screwdriver / 1 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly Exploded view for installation Figure 5.21 Disassembling/assembling bar code reader M3×10 hexagon socket head cap screws and spring washer and big flat washer Small bracket for bar code reader Bar code reader How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carosuel cover, and remove the left panel assembly, front cover assembly, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the sample probe from colliding during the deatching. 3 Remove the connector of the bar code reader. 4 Remove the 2 M3×10 hexagon socket head cap screws, the spring washer, and the big flat washer, and remove the bar code reader. 5 Install a new bar code reader into a small bracket mounting position of the bar code reader. Pre-tighten the screw, and plug the cable of the bar code reader. 6 Power on the bar code reader, re-align the bar code scan position, and use an adjustment tool of the bar code scan light of the BA24-J03 BA24 to perform alignment according to steps described on the software. After the light of the bar code reader is adjusted, tighten the retaining screw. 7 Assemble and reset the panel removed in step 2. Alignment and confirmation See 8.8.3 to align the bar code scan position. BS-240 Service Manual 5-39 5 Modules and Units 5.4 Sample/reagent carousel assembly 5.4.11 Replacing Reagent Carousel Antifogging Heating Assembly When to do Replace the assembly when the glass of the antifogging assembly is damaged or the assembly fails. Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Rubber gloves for medical use / 1 pair M3×10 hexagon socket head cap screws and spring washer and big flat washer Small bracket for bar code reader Reagent carousel antifogging heater assembly Dust shield and sponge M2×8 cross recessed pan head screw Antifogging heater Temperature switch Anti-fogging mounting plate M3×8 cross recessed pan head screw Heat conductive plate Bar code reader Switch anchor plate M3×6 cross recessed pan head screw M3×12 hexagon socket cap head screw Glass window Rubber cushion How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carosuel cover, and remove the left panel assembly, front cover assembly, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the sample probe from colliding during the deatching. 3 Remove the 2 M3×10 hexagon socket head cap screws, the spring washer, and the big flat washer, and remove the bar code reader. 4 Remove the 4 M3×6 cross recessed pan head screws, and remove the dust shield with sponge. 5 Remove the connectors of the temperature switch and the antifogging heater, unscrew the 4 M3×8 cross pan head screws, and remove the antifogging heating assembly of the reagent carousel. 6 5-40 Remove 1 M3×12 hexagon socket head cap screws, and remove the anchor plate of the switch, the temperature protection switch, and the antifogging heater. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.4 Sample/reagent carousel assembly 7 Remove 4 M2×8 cross recessed pan head screws, remove the thermal conductive aluminum plate and the damaged glass pane, and clear the broken glass in the slot of the mounting plate. Replace the rubber pad when necessary. 8 Place a new glass pane onto the silicone pad in the slot of the mounting plate of the antifogging apparatus. Ensure that the glass is not cracked or dirty. 9 Use 4 M2×8 cross recessed pan head screws to fasten the thermal conductive aluminum plate into the mounting position. 10 Apply thermal paste onto the fitting surface between the antifogging heater (if the heater is damaged, use a new antifogging heater) and the thermal conductive aluminum plate, and place the heater in the mounting position. 11 Apply thermal paste onto the fitting surface between the temperature switch (if the temperature switch is damaged, use a new temperature switch) and the antifogging heater, and place switch in a middle position. Place the anchor plate of the switch in the mounting position, and fasten it with 1 M3×12 hexagon socket head cap screw. 12 Use 4 M3×8 cross recessed pan head screws to install the reagent carousel antifogging assembly into the mounting position of the reagent refrigeration compartment. 13 Plug the cable of the temperature switch and the antifogging heater, use 4 M3×6 cross pan head screws to fasten the dust shield with sponge onto the mounting plate of the antifogging apparatus. 14 Install the bar code reader into a small bracket of the bar code reader, and pretighten the screw (plug the cable of the bar code reader). 15 Power on the bar code reader, re-align the bar code scan position, and use an adjustment tool of the bar code scan light of the BA24-J03 BA24 to perform alignment according to steps described on the software. After the light of the bar code reader is adjusted, tighten the retaining screw. 16 Assemble and reset the panel removed in step 2. Alignment and confirmation See 8.8.3 to align the bar code scan position. BS-240 Service Manual 5-41 5 Modules and Units 5.5 Sample Probe Movement Assembly 5.5 Sample Probe Movement Assembly 5.5.1 Main functions The sample probe movement assembly includes a sample probe drive assembly, a sample probe arm assembly, and a sample probe assembly. The sample probe movement assembly can implement horizontal rotation movements and vertical movements; and have the following functions: fluid level detection, clog detection, and vertical bump detection; and have other functions such as mechanical limit stop, and power failure selflock. Horizontal direction: The sample probe center can pass through the wash well, the sample reagent carousel aspiration position, the reaction carousel/ISE dispensing position->wash well->sample/reagent carousel aspiration position-> reaction carousel dispensing position->wash well position, and meet the movement stroke requirements. Vertical direction: The sample probe can move vertically at the reagent bottle, sample container, reaction cuvette, wash well, and ISE dispensing port, and meet the movement stroke requirements. 5-42 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.5 Sample Probe Movement Assembly Figure 5.22 Location of sample probe movement assembly in the instrument Sample Probe Movement Assembly Mixer Movement Assembly Shielding Cover Sample/reagent carousel cover Sample/reagent panel Reagent bar code assembly Wash station desk panel Reaction carousel panel Lamp panel Frame Assembly Reaction Carousel Assembly Left panel assembly BS-240 Service Manual Sample/reagent carousel assembly Front cover assembly Reagent refrigeration board and the bracket 5-43 5 Modules and Units 5.5 Sample Probe Movement Assembly 5.5.2 Assembly Locations and FRU Details Figure 5.23 Sample probe movement assembly 4 5 6 3 7 2 8 1 9 Table 5.5 List of FRU materials No Part Name Remark 1 009-002204-00 Correlative photocoupler (S) FRU 2 115-037085-00 BA24 HbA1c sample probe assembly FRU 3 033-000108-00 Anti-collision spring / 4 115-037552-00 Reagent preheating assembly FRU 5 051-002479-00 Level sense board PCBA FRU 6 041-009179-00 Hardened power screw (M3×6) / 7 BA30-10-06619 Synchronous belt B123MXL6.4 FRU 8 115-016625-00 3-probe drive horizontal motor assembly FRU 9 115-036662-00 Probe drive assembly FRU 10 044-000263-00 130mm arm base FRU 11 043-002344-00 130mm arm cover FRU 12 115-016462-00 Guide rod and anti-collision spring FRU 115-005246-00 42 step motor 1 with belt pulley (horizontal) FRU 115-005247-00 59 step motor 1 with belt pulley (vertical) FRU 115-015298-00 3-probe drive vertical motor assembly FRU 13 14 15 5-44 FRU No. or Part No. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.5 Sample Probe Movement Assembly 5.5.3 Replacing Horizontal Movement Photocoupler and Vertical Movement Photocoupler When to do Replace the horizontal movement photocoupler and the vertical movement photocoupler when they fail. Tools Name Code Quantity Cross screwdriver / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.24 Disassembling/assembling horizontal movement photocoupler and vertical movement photocoupler M3×6 cross recessed pan head screw M3×6 cross recessed pan head screw Horizontal sensor bracket Horizontal movement photocoupler Vertical movement photocoupler M3×6 cross recessed pan head screw How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover and remove the lamp panel, the reaction carousel panel, and the wash panel. Protect the sample probe from colliding during the disassembling. BS-240 Service Manual 5-45 5 Modules and Units 5.5 Sample Probe Movement Assembly 3 When replacing the vertical movement photocoupler, remove the photocoupler cable, remove 1 M3×6 cross recessed pan head screw to take out the vertical movement photcoupler, use 1 M3×6 cross recessed pan head screw to fasten a new photocoupler, and plug the cable. 4 When replacing the horizontal movement photocoupler, remove the photocoupler cable, remove 1 M3×6 cross recessed pan head screw to take out the horizontal sensor bracket with the photocoupler, remove 1 M3×6 cross recessed pan head screw to take out the horizontal movement photcoupler, use 1 M3×6 cross recessed pan head screw to fasten a new photocoupler onto the horizontal sensor bracket, use 1 M3×6 cross recessed pan head screw to fasten the horizontal sensor bracket in the mounting position, and plug the photocoupler cable. 5 Install the panel that is disassembled in step 2. Alignment and confirmation N/A 5.5.4 Replacing Sample Probe Assembly, Anti-collision Spring and Power Screw When to do Replace them when the sample probe assembly is damaged or the anti-collision spring fails or the power screw has poor appearance. Tools Name 5-46 Code Quantity Cross screwdriver / 1 Flathead screwdriver / 1 Rubber gloves for medical use / 1 pair BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.5 Sample Probe Movement Assembly Exploded view for installation Figure 5.25 Disassembling/assembling sample probe assembly, anti-collision spring and power screw 130mm arm cover Spring guide post Anti-collision spring Probe assembly Hardened power screw (M3×6) How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover. 3 Remove 2 hardened power screws (M3×6) to take out the 130mm arm cover. The powder screw may be replaced directly. 4 Use a flathead screwdriver to loosen the guide rod and take out the guide rod and the anti-collision spring. The anti-collision spring may be replaced directly with a new spring. 5 Remove the anti-collision detection cable connector, and clean the tube connector. 6 Remove the sample probe assembly, replace it with a new sample probe assembly, plug the cable connector, and clean the tube connector. 7 Assemble and reset the assembly in the reversed order of steps 2~4. Alignment and confirmation N/A BS-240 Service Manual 5-47 5 Modules and Units 5.5 Sample Probe Movement Assembly 5.5.5 Replacing Reagent Preheating Assembly When to do Replace the reagent preheating assembly when it fails. Tools Name Code Quantity Cross screwdriver / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.26 Disassembling/assembling reagent preheating assembly 130mm arm cover Reagent preheating assembly Insulating plate M3×8 cross recessed pan head screw Hardened power screw (M3×6) How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover. 3 Remove 2 hardened power screws (M3×6) to take out the 130mm arm cover. 4 Remove the reagent preheating assembly cable, remove the wash tube connector from the sample probe, and remove the quick connector. 5 Remove 2 M3×8 cross recessed pan head screws, remove 2 insulation pads, and lay flat the reagent preheating assembly. Disassemble the heating cover of the reagent preheating assembly, and take out the wash tube. 5-48 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.5 Sample Probe Movement Assembly 6 Prepare a new reagent preheating assembly, disassemble the heating cover of the reagent preheating assembly, and assemble the heating cover after threading the wash tube. 7 Place 2 insulation pads in the mounting position, and use two M3×8 cross recessed pan head screws to pass through the insulation pads, and fasten the reagent preheating assembly. 8 Tidy up the cable, interconnect the cable, tidy up the wash tube, reinstall the terminals and the quick connector, and screw the quick connector into the assembly. 9 Use the sealing adhesive to seal the outlet and inlet of the wash tube of the reagent preheating assembly, and the cable hole ends. 10 Assemble and reset the assembly in the reversed order of steps 2~3. Alignment and confirmation N/A 5.5.6 Replacing Level Sense Board When to do Replace the level sense board when necessary. Tools Name BS-240 Service Manual Code Quantity Cross screwdriver / 1 Rubber gloves for medical use / 1 pair 5-49 5 Modules and Units 5.5 Sample Probe Movement Assembly Exploded view for installation Figure 5.27 Disassembling/assembling level sense board 130mm arm cover Level sense board M3×6 cross recessed pan head screw Hardened power screw (M3×6) How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover. 3 Remove 2 hardened power screws (M3×6) to take out the 130mm arm cover. 4 Remove the connector on the level sense board. 5 Remove 2 M3×6 cross recessed pan head screws, and remove the level sense board. 6 Replace the board with a new level sense board, and install and reset it in the reversed order of the steps stated above. Alignment and confirmation N/A 5.5.7 Replacing Horizontal Rotating Synchronous Belt When to do Replace the horizontal rotating synchronous belt when it fails. Tools Name 5-50 Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Rubber gloves for medical use / 1 pair BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.5 Sample Probe Movement Assembly Exploded view for installation Figure 5.28 Disassembling/assembling horizontal rotating synchronous belt Horizontal rotary synchronous belt M3×8 hexagon socket head cap screw How to do Steps of replacing the horizontal rotating synchronous belt: 1 Switch off the power of the analyzer. 2 Open the shielding cover and remove the lamp panel, the reaction carousel panel, and the wash panel. Protect the sample probe from colliding during the disassembling. 3 Draw an arc along the flat washers under the two M3×8 hexagon socket head cap screws that fasten the outside of the motor. 4 Loosen 4 M3×8 hexagon socket head cap screws that fasten the motor, and remove the synchronous belt. 5 Sheathe a new synchronous belt slowly from the tip into two pulleys. 6 Move the motor. When the flat washers under the two outside screws basically overlap the arc drawn in step 3, tighten the 4 screws to fasten the motor. 7 Reset the panel disassembled in step 2. Alignment and confirmation N/A 5.5.8 Replacing Horizontal Motor Assembly When to do Replace the horizontal motor assembly when it fails. BS-240 Service Manual 5-51 5 Modules and Units 5.5 Sample Probe Movement Assembly Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.29 Horizontal motor assembly Horizontal rotary synchronous belt M3×8 hexagon socket head cap screw How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover and remove the lamp panel, the reaction carousel panel, and the wash panel. Protect the sample probe from colliding during the disassembling. 3 Draw an arc along the flat washers under the two M3×8 hexagon socket head cap screws that fasten the outside of the motor. 4 Remove the motor cable connector, remove 4 M3×8 hexagon socket head cap screws and the spring/flat washers that are used to fasten the motor, and remove the horizontal motor assembly. 5 Use 4 M3×8 hexagon socket head cap screws and spring/flat washers to pre-tighten a new horizontal motor, and move the motor until the flat washers under the two outside screws basically overlap the arc drawn in step 3. Tighten the 4 screws to fasten the motor, and plug the motor cable connector. 6 Reset the panel disassembled in step 2. Alignment and confirmation N/A 5-52 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.5 Sample Probe Movement Assembly 5.5.9 Replacing Sample Probe Drive Assembly When to do Replace the sample probe drive assembly when it fails. Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Diagonal pliers / 1 Exploded view for installation Figure 5.30 Horizontal motor assembly 130mm arm cover Sample probe arm assembly M3×10 hexagon socket head cap screws and spring washer Hardened power screw (M3×6) Probe drive assembly M6×16 hexagon socket head cap screws and spring/flat washer BS-240 Service Manual 5-53 5 Modules and Units 5.5 Sample Probe Movement Assembly How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carousel cover, and remove the left panel assembly, front cover assembly, rear panel, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the sample probe from colliding during the disassembling. 3 Remove the cable and wash tube that pass through the precision shaft, remove all the cables and tube terminals, and take out the braided hose. 4 Remove vertical and horizontal motor connectors, and remove 2 photocoupling connectors. 5 Remove 2 hardened power screws (M3×6) to take out the arm cover. 6 Remove 2 M3×10 hexagon socket head cap screws and spring washers to take out the sample probe arm assembly, and move the cable out of the precision shaft. Prevent probe collision. 7 Remove 3 M6×16 hexagon socket head cap screws and spring/flat washers, and take out the sample probe drive assembly from the instrument. 8 Place a new sample probe drive assembly into the mounting position, and use 3 M6×16 hexagon socket head cap screws and spring/flat washers to fasten the assembly. 9 Thread the cable and tube on the sample probe arm assembly from the top of the precision shaft, and use 2 M3×10 hexagon socket head cap screws and spring washers to pre-tighten the arm assembly onto the precision shaft. 10 Sheathe the braided hose into the cable and the tube, reinstall connectors of the cable, and insert the cable into the corresponding slot. Screw the tube mounting terminals and the quick connector into the corresponding port. 11 Connect the motor cable and the photocoupler cable. 12 See sections 8.5.1~8.5.9 to align positions from the sample probe to the reaction carousel, wash well, sample/reagent carousel and ISE, tighten 2 M3×10 hexagon socket head cap screws to fasten the arm assembly, and use 2 hardened powder screws (M3×6) to install the arm cover. 13 Reassemble and reset the parts disassembled in step 2. Alignment and confirmation See sections 8.5.1~8.5.9 to align positions from the sample probe to the reaction carousel, wash well, sample/reagent carousel, and ISE. 5-54 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.6 Mixer Movement Assembly 5.6 Mixer Movement Assembly 5.6.1 Main functions The mixer movement assembly consists of a mixer drive assembly and an arm assembly. The assembly can implement horizontal rotating movements and vertical movements, and can stir the reaction mixture when the sample and the reagent are successively added. Horizontal direction: The mixer center can pass through the wash well and the reaction carousel mixing position, and meet movement stroke requirements. Vertical direction: The mixer can perform vertical movements in the reaction cuvette and the wash well, and meet the movement stroke requirements. Figure 5.31 Location of mixer movement assembly in the instrument Sample Probe Movement Assembly Shielding Cover Sample/reagent carousel cover Sample/reagent panel Reagent bar code assembly Mixer Movement Assembly Wash station desk panel Reaction carousel panel Lamp panel Frame Assembly Reaction Carousel Assembly Left panel assembly BS-240 Service Manual Sample/reagent carousel assembly Front cover assembly Reagent refrigeration board and the bracket 5-55 5 Modules and Units 5.6 Mixer Movement Assembly 5.6.2 Assembly Locations and FRU Details The mixer movement assembly is located right above the instrument. Figure 5.32 Mixer movement assembly 4 3 5 6 2 1 7 Table 5.6 List of materials No 5-56 FRU No. or Part No. Part Name Remark 1 009-002204-00 Correlative photocoupler (S) FRU 2 BA30-10-06619 Synchronous belt B123MXL6.4 FRU 3 115-016625-00 3-probe drive motor assembly FRU 4 BA30-10-15052 DC motor WRF-130CH FRU 5 BA31-20-41651 Mixer FRU 6 041-009179-00 Hardened power screw (M3×6) / 7 115-036665-00 Mixer drive assembly FRU 8 044-000333-00 65mm mixer arm base FRU 9 043-003064-00 65mm mixer arm cover FRU 10 BA31-20-41651+BA30-20-06774 Mixer assembly FRU 11 115-015298-00 3-probe drive vertical motor assembly FRU 12 115-005246-00 42 step motor 1 with belt pulley (horizontal) FRU 13 115-005247-00 59 step motor 1 with belt pulley (vertical) FRU horizontal BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.6 Mixer Movement Assembly 5.6.3 Replacing Horizontal Movement Photocoupler and Vertical Movement Photocoupler When to do Replace the horizontal movement photocoupler and the vertical movement photocoupler when they fail. Tools Name Code Quantity Cross screwdriver (long, short) / 1 each Adjustable wrench / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.33 Exploded view for installing horizontal movement photocoupler and vertical movement photocoupler Wash well support shaft Mixer Movement Assembly M3×6 cross recessed pan head screw Horizontal movement photocoupler Horizontal sensor bracket M3×6 cross recessed pan head screw Vertical movement photocoupler M3×6 cross recessed pan head screw Replacement steps: 1 Switch off the power of the analyzer. 2 Open the shielding cover and remove the lamp panel, the reaction carousel panel, and the wash panel. Protect the mixer from colliding during the disassembling. BS-240 Service Manual 5-57 5 Modules and Units 5.6 Mixer Movement Assembly 3 When replacing the vertical movement photocoupler, remove the wash well support shaft, remove the photocoupler cable, use a short crosshead screwdriver to take out 1 M3×6 cross recessed pan head screw and take out the vertical movement photcoupler, use 1 M3×6 cross recessed pan head screw to fasten a new photocoupler, and plug the cable. Install the wash well support shaft. 4 When replacing the horizontal movement photocoupler, remove the photocoupler cable, remove 1 M3×6 cross recessed pan head screw to take out the horizontal sensor bracket with the photocoupler, remove 1 M3×6 cross recessed pan head screw to take out the horizontal movement photcoupler, use 1 M3×6 cross recessed pan head screw to fasten a new photocoupler onto the horizontal sensor bracket, use 1 M3×6 cross recessed pan head screw to fasten the horizontal sensor bracket in the mounting position, and plug the photocoupler cable. 5 Install the panel that is disassembled in step 2. Alignment and confirmation N/A 5.6.4 Replacing horizontal rotating synchronous belt When to do Replace the horizontal rotating synchronous belt when it fails. Tools Name 5-58 Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Rubber gloves for medical use / 1 pair BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.6 Mixer Movement Assembly Exploded view for installation Figure 5.34 Installing horizontal rotating synchronous belt Horizontal rotary synchronous belt M3×8 hexagon socket head cap screw Replacement steps: 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carosuel cover, and remove lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the mixer from colliding during the disassembling. 3 Draw an arc along the flat washers under the two M3×8 hexagon socket head cap screws that fasten the outside of the motor. 4 Loosen 4 M3×8 hexagon socket head cap screws that fasten the motor, and remove the synchronous belt. 5 Sheathe a new synchronous belt slowly from the tip into two pulleys. 6 Move the motor. When the flat washers under the two outside screws basically overlap the arc drawn in step 3, tighten the 4 screws to fasten the motor. 7 Reset the panel disassembled in step 2. Alignment and confirmation N/A 5.6.5 Replacing Horizontal Motor Assembly When to do Replace the horizontal motor assembly when it fails. BS-240 Service Manual 5-59 5 Modules and Units 5.6 Mixer Movement Assembly Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.35 Exploded view for installing horizontal motor Horizontal rotary synchronous belt M3×8 hexagon socket head cap screws and spring/flat washer Horizontal motor assembly Replacement steps: 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carousel cover, and remove lamp panel, reaction carousel panel, wash panel, and sample/reagent panel. Protect the mixer from colliding during the disassembling. 3 Draw an arc along the flat washers under the two M3×8 hexagon socket head cap screws that fasten the outside of the motor. 4 Remove the motor cable connector, remove 4 M3×8 hexagon socket head cap screws and the spring/flat washers that are used to fasten the motor, and remove the horizontal motor assembly. 5 Use 4 M3×8 hexagon socket head cap screws and spring/flat washers to pre-tighten a new horizontal motor, and move the motor until the flat washers under the two outside screws basically overlap the arc drawn in step 3. Tighten the 4 screws to fasten the motor, and plug the motor cable connector. 6 5-60 Reset the panel disassembled in step 2. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.6 Mixer Movement Assembly Alignment and confirmation N/A 5.6.6 Replacing DC Motor, Mixer and Powder Screw When to do Replace them when the DC motor fails, or the mixer is bent, or the power screw has poor appearance. Tools Name Code Quantity Cross screwdriver (big and small) / 1 each Rubber gloves for medical use / 1 pair Exploded view for installation Figure 5.36 Exploded view for installing DC motor, mixer and powder screw Mixer arm cover DC motor Hardened power screw (M3×6) M2×4 cross recessed pan head screw and flat washer Retaining nut Mixer Replacement steps: 1 Switch off the power of the analyzer. 2 Open the shielding cover and remove the lamp panel. When necessary, remove the left panel assembly and the front panel assembly. 3 Remove 2 hardened power screws (M3×6) to take out the mixer arm cover. The powder screw may be replaced directly. 4 Loosen the retaining nut and remove the mixer. When repairing the mixer, replace the mixer with a new mixer, and insert the mixer into the motor shaft and tighten the retaining nut. BS-240 Service Manual 5-61 5 Modules and Units 5.6 Mixer Movement Assembly 5 Remove the motor cable, remove 2 M2×4 cross countersunk head screw and flat washers, take out the DC motor, replace the motor with a new DC motor, and use 2 M2×4 cross countersunk head screws and flat washers to pre-tighten the motor. Plug the motor cable. 6 Install the mixer, tighten the retaining nut and install the mixer arm cover. 7 Install the cover that is disassembled in step 2. Alignment and confirmation See 8.6.1 to align horizontal positions from mixer to reaction carousel 5.6.7 Replacement of Mixer Drive Assembly When to do Replace the mixer drive assembly when it fails. Tools Name 5-62 Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Diagonal pliers / 1 Rubber gloves for medical use / 1 pair BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.6 Mixer Movement Assembly Exploded view for installation Figure 5.37 Exploded view of mixer drive assembly Mixer arm cover Mixer arm assembly M3×10 hexagon socket head cap screws and spring washer Hardened power screw (M3×6) Mixer drive assembly M6×16 hexagon socket head cap screws and spring/flat washer Replacement steps: 1 Switch off the power of the analyzer. 2 Open the shielding cover, remove the sample/reagent carousel cover, and remove the left panel assembly, front cover assembly, rear panel, lamp panel, reaction carosuel panel, wash panel, and sample/reagent panel. Protect the mixer from colliding during the disassembling. 3 Remove the cable that passes through the precision shaft, remove all the cables and terminals, and take out the braided hose. 4 Remove vertical and horizontal motor connectors, and remove 2 photocoupling connectors. 5 Remove 2 hardened power screws (M3×6) to take out the arm cover. 6 Remove 2 M3×10 hexagon socket head cap screws and spring washers to take out the mixer arm assembly, and move the cable out of the precision shaft. Prevent mixer collision. 7 Remove 3 M6×16 hexagon socket head cap screws and spring/flat washers, and take out the mixer drive assembly from the instrument. 8 Place a new mixer drive assembly into the mounting position, and use 3 M6×16 hexagon socket head cap screws and spring/flat washers to fasten the assembly. BS-240 Service Manual 5-63 5 Modules and Units 5.6 Mixer Movement Assembly 9 Thread the cable on the mixer arm assembly from the top of the precision shaft, and use 2 M3×10 hexagon socket head cap screws and spring washers to pre-tighten the arm assembly onto the precision shaft. 10 Sheathe the braided hose into the cable, reinstall connectors of the cable, and insert the cable into the corresponding slot. 11 Connect the motor cable and the photocoupler cable. 12 See sections 8.6.1~8.6.3 to align positions from the mixer to the reaction carousel and wash well, tighten 2 M3×10 hexagon socket head cap screws to fasten the arm assembly, and use 2 hardened powder screws (M3×6) to install the arm cover. 13 Reassemble and reset the parts disassembled in step 2. Alignment and confirmation See 8.6.1~8.6.3 to align positions from mixer to reaction carousel and wash well. 5-64 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.7 Wash Assembly 5.7 Wash Assembly 5.7.1 Main functions The wash assembly, located in the rear right of the whole unit, consists of the wash probe assembly and wash probe drive assembly. It provides 4-phase wash for non-disposable reaction cuvettes so that they can be used repeatedly without influencing the test effects. The first 3 phases are wash phase probe assemblies. The wash probes of the first 1 to 3 phases have dispense and aspiration functions. The 4th phase is a wipe phase probe assembly. Figure 5.38 Location of wash assembly Sample/reagent carousel assembly Frame Assembly BS-240 Service Manual Sample Probe Movement Assembly Mixer Movement Assembly Auto Wash Assembly Reaction Carousel Assembly Optical assembly 5-65 5 Modules and Units 5.7 Wash Assembly 5.7.2 Locations and FRU Details Figure 5.39 Wash assembly Knurled screw Wash probe drive assembly Zero position photocoupler Table 5.7 List of materials No Part Name Remark 1 115-036353-00 Wash probe assembly FRU 2 051-001147-00 BA48 wash station photocoupler PCBA FRU 3 115-036352-00 Wash probe assembly FRU 4 009-002204-00 Photocoupler cable (S) FRU / Zero position photocoupler BA38-30-88154 Wash phase assembly probe FRU BA40-30-61934 Wipe phase assembly probe FRU 041-001845-00 Wipe block 5 6 7 5-66 FRU No. or Part No. drive FRU BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.7 Wash Assembly 5.7.3 Replacing Wash Probe Assembly When to do Replace the wash probes when they are bent or damaged. Tools Name Cross screwdriver Code / Quantity 1 Exploded view for installation Figure 5.40 Wash probe assembly Cable anchor plate BA48 wash station photocoupler M2.5×4 cross recessed pan head screw How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover assembly, unscrew 8 M4×8 cross pan head combination screws to remove the rear panel. 3 Unplug the cable connector of the BA48 wash station photocoupler. 4 Unplug the tubes and cables on the wash probe assembly, and manually loosen the knurled screw to remove the wash probe assembly. 5 Install the new wash probe assembly, manually tighten the knurled screw, and restore the tubes and cables according to their numbers. 6 Restore the components in the reversed order. BS-240 Service Manual 5-67 5 Modules and Units 5.7 Wash Assembly NOTE: Reinstall the tubes and cables on the wash probe assembly according to the number marking. Alignment and confirmation Check the wash probes' vertical position in reaction cuvettes according to 8.7.3. 5.7.4 Replacing Wash Station Photocoupler When to do Replace the wash station photocoupler when it is damaged. Tools Name Cross screwdriver Code / Quantity 1 Exploded view for installation Figure 5.41 Wash station photocoupler Cable anchor plate M2.5×4 cross BA48 wash station recessed pan head screw photocoupler How to do 1 5-68 Switch off the power of the analyzer. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.7 Wash Assembly 2 Open the shielding cover assembly, unscrew 8 M4×8 cross pan head combination screws to remove the rear panel. 3 Unplug the cable connector of the wash station photocoupler. 4 Unplug the tubes and cables on the wash probe assembly. 5 Manually loosen the knurled screw to remove the wash probe assembly. 6 Loosen 2 M2.5×8 cross countersunk head screws on the anchor plate and then remove the anchor plate. 7 Loosen 1 M2.5×4 cross pan head screw on the wash station photocoupler, replace it with a new one, and tighten 1 M2.5×4 cross pan head screw. 8 Restore the components in the reversed order. NOTE: Reinstall the tubes and cables on the wash probe assembly according to the number marking. Alignment and confirmation N/A 5.7.5 Replacing Wash Probe Drive Assembly When to do Replace the wash probe drive assembly when it is damaged. Tools Name BS-240 Service Manual Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 5-69 5 Modules and Units 5.7 Wash Assembly Exploded view for installation Figure 5.42 Wash probe drive assembly Knurled screw Position adjusting plate M4×16 hexagon socket head cap screw + spring washer Zero position photocoupler M3×8 cross recessed pan head screw M4×12 hexagon socket head cap screw + spring washer Motor support block How to do 1 Switch off the power of the analyzer. 2 Open the shielding cover, loosen the knurled screws on the wash probe assembly, remove the wash probe assembly, and place the probe properly against bending. Remote the sample probe to the reagent aspirating port position, remote the mixer to the wash well position, unscrew the M4X8 cross recessed pan head padded screw, remove the BA24 reaction carousel panel (basic part), lamp panel, wash panel, unscrew the 3 M4X8 cross recessed pan head screws for fastening the right panel assembly, and remove the right panel assembly. 3 Unplug the zero position photocoupler and the motor, loosen 4 M4×12 hexagon socket head cap screws with spring washer on the motor supporting block, replace the wash probe drive assembly with a new one, and then fix it with 4 M4×12 hexagon socket head cap screws with spring washer. Do not tighten the 2 M4×16 hexagon socket head cap screws with spring washer on the wash probe drive assembly, till you finish the alignment. 4 5-70 Restore the components in the reversed order. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.7 Wash Assembly Alignment and confirmation Check the wash probes' vertical and horizontal positions in reaction cuvettes according to 8.7.2~8.7.3. 5.7.6 Replacing Zero Position Photocoupler When to do Replace the zero position photocoupler on the wash probe drive assembly when it is damaged. Tools Name Code Quantity Cross screwdriver / 1 Hexagon wrench / 1 Exploded view for installation Figure 5.43 Zero position photocoupler on the wash probe drive assembly Knurled screw Position adjusting plate M4×16 hexagon socket head cap screw + spring washer Zero position photocoupler M3×8 cross recessed pan head screw M4×12 hexagon socket head cap screw + spring washer Motor support block BS-240 Service Manual 5-71 5 Modules and Units 5.7 Wash Assembly How to do 1 Switch off the power of the analyzer. 2 Unscrew the 8 M4×8 cross pan head combination screws on the rear panel. Remove the syringe drive assembly when removing the rear panel. 3 Unplug the connector of the zero position photocoupler. 4 Loosen 1 M3×8 cross pan head screw on the wash probe drive assembly, replace the zero position photocoupler with a new one, and then fix it with 1 M3×8 cross pan head screw. 5 Restore the components in the reversed order. Alignment and confirmation Adjust the wash probes' vertical position in reaction cuvettes according to 8.7.3. 5-72 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.8 Syringe Assembly 5.8 Syringe Assembly 5.8.1 Main functions The syringe assembly includes 1 sample syringe assembly and 2 wash syringe assemblies. 1 of the wash syringe assemblies is used to inject wash solution, and is a 2.5ml syringe assembly; and the other is used to inject deionized water, and is a hematology 10ml syringe assembly. 5.8.2 Assembly Locations and FRU Details Figure 5.44 Syringe assembly Frame Syringe bracket weld Syringe module Syringe left bracket 2.5ml syringe assembly Syringe left bracket 10 ml screw rod driving syringe assembly Power supply assembly Syringe lower bracket Table 5.8 List of materials No 1 BS-240 Service Manual FRU No. or Part No. Part Name Remark 115-011901-00 10 ml screw rod driving syringe assembly FRU 2 115-031800-00 2.5ml syringe assembly FRU 3 2800-21-28878 Motor position sensor assembly FRU 4 115-036403-00 BA25 syringe assembly / 5-73 5 Modules and Units 5.8 Syringe Assembly 5.8.3 Replacing 2.5ml Syringe and 10ml Syringe When to do Replace the syringe when the syringe is abnormal. Tools Name Code Quantity cross screwdriver / 1 Hexagon wrench / 1 Exploded view for installation See syringe assembly. Replacement steps: 1 Switch off the power supply of the whole unit. 2 Pull out the tube and lines on the syringe, and use a hexagon socket wrench for loosening. Fasten the 2.5ml syringe to the 4 hexagon screws of the left bracket and lower bracket of the syringe, so that the 2.5ml syringe assembly can be taken out. 3 Pull out the tube and lines on the syringe, and use a hexagon socket wrench for loosening. Fasten the 10ml screw rod driving syringe to the 4 hexagon screws of the right bracket and lower bracket of the syringe, so that the 10ml screw rod driving syringe assembly can be taken out. 4 After replacing the syringe with a new syringe, restore the assembly in the reversed order. 5.8.4 Replacing Syringe Refer to Section 9.3.2. 5-74 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.9 Photoelectric Unit 5.9 Photoelectric Unit 5.9.1 Functions and Parameters The BS-240 has inherited the mature filter light-splitting technology of Mindray's BS series. The photoelectric detection unit uses a light source to illuminate directly. After passing through the interference filter, composite light becomes monochromatic light of a specific bandwidth. After being absorbed by a solution, the monochromatic beam is received by a photodetector. The photodetector converts the optical signal with light intensity into an electric signal. A microprocessing unit calculates the concentration of the solution by calculating the change of the light intensity. Multiple monochromatic wavelengths required in a photometric system are implemented by using a filter wheel apparatus. During measurement, filters are sequentially shifted into the optical path to implement colorimetric measurement of each wavelength. Technical Parameters Optical system method: filter wheel forward-splitting system, optical-to-electric conversion of silicon photodiode. Wavelength range: 8 wavelengths: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, and 670nm Wavelength accuracy: ±2nm Light source index: 12V/20W tungsten-halogen lamp Minimum reaction volume: 100μl Absorbance range: 0-4A, 10mm light pathlength Optical pathlength of reaction cuvette: 5mm Number of cuvettes: 8 segments, including 40 cuvettes in total Cuvette cavity dimensions: 5mm×5mm×29.5mm Cuvette material: plastic Terms of Optics AD value: the value converted from photoelectric signal (voltage) through AD converter. Water blank: the AD value of a cuvette measured when the wash station dispenses water in phase 6. Water blank is the base point for calculation absorbance, that is, the 0 point of absorbance. Cuvette blank out of range: means the water blank of phase 6 is less than the light intensity low limit. This alarm is used to monitor the energy level of the optical system to ensure normal signal-to-noise ratio. BS-240 Service Manual 5-75 5 Modules and Units 5.9 Photoelectric Unit Cuvette blank out of range (10X): means the relative change of continuous phase-6 water blanks for 10 cuvettes is greater than 3% comparing with the history data. This alarm is used to prevent light fluctuation or erroneous result due to cuvette overflowing and to remind the operator of this phenomenon. 5.9.2 Composition and Structure of Optical Assembly The BS-240 optical assembly uses a filter wheel plus a PD to for photometric measurement. See the figure below. The light source illuminates directly, and the beam is converged by the lens 1 and then split by the filter. A monochromatic beam is shaped by the lens 2 and then passes through the cuvette. After being absorbed by the solution, the beam is converged by the lens 3. Finally, the optical signal is received by the photodiode and converted into an electric signal for outputting. The whole optical path is from the reaction carousel inwards. Figure 5.45 Optical path Light source Lens 1 Filter Cuvettes Lens 3 Photodiode Lens 2 5-76 Optical base BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.9 Photoelectric Unit Figure 5.46 Location of optical assembly Optical assembly The optical assembly consists of a light source assembly and a measurement assembly. Table 5.9 List of materials No BS-240 Service Manual FRU No. or Part No. Part Name Remar ks 1 115-038466-00 20W lamp assembly (FRU) FRU 2 041-001143-00 Lamp base retaining screw FRU 3 BA10-20-78211 Patch cord for lamp housing fan FRU 4 801-BA10-00099-00 Preamplifier board PCBA(FRU/C) FRU 5 115-036571-00 Cuvette (Surface processed) FRU 6 801-BA10-00041-00 Filter 340 (C) FRU 7 801-BA10-00042-00 Filter 405 (C) FRU 8 801-BA10-00043-00 Filter 450 (C) FRU 9 801-BA10-00044-00 Filter 510 (C) FRU 10 801-BA10-00045-00 Filter 546 (C) FRU 11 801-BA10-00046-00 Filter 578 (C) FRU 12 801-BA10-00047-00 Filter 630 (C) FRU 13 801-BA10-00048-00 Filter 670 (C) FRU 14 801-BA10-00004-00 AD collection board (C) FRU 5-77 5 Modules and Units 5.9 Photoelectric Unit 5.9.3 Replacing Lamp When to do Replace the lamp when an alarm is triggered indicating that the lamp has insufficient intensity, or not turned on, or has been used for over 2000 hours, or has cuvette blank out of range (10X). Tools Cross screwdriver Exploded view for installation Figure 5.47 Exploded view of lamp mounting position Optical assembly Lamp assembly Wiring terminal nut (screwed manually) Lamp retaining screw NOTE: Replace the lamp according to the instructions on the screen. Please wait for at least 5 minutes (if the lamp fan works properly) after turning off the lamp to avoid injury. Do not touch the bulb when installing the new lamp. In case of touching, please clean the bulb with fiber cloth or ethanol-moistened gauze. How to do 5-78 1 Select Utility -> Maintenance -> Maintenance -> Biochemistry Maintenance, and select Replace Lamp. 2 Follow the instructions prompted by the software and wait for 10 minutes until the lamp cools down. 3 Use a cross screwdriver to loosen the screws of the lamp panel, and remove the panel. 4 Unscrew the cable connectors and the lamp base retaining screws with hands, to remove the lamp. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.9 Photoelectric Unit 5 Install the lamp and panel in a reversed order. Before installing the panel, ensure that the lamp has been installed properly in the right place and there is no gap between the lamp and the lamp base. 6 Select "Continue" and wait for 5 minutes for the new lamp to get steady. Select "Photometer Check" to check if the lamp can work normally. Alignment and confirmation Refer to the alignment section. 5.9.4 Disassembling/Assembling Optical Assembly When to do Remove the optical assembly before replacing the filter, preamplifier board, and lamp radiating fan. Tools Cross screwdriver and hexagon wrench Exploded view for installation Figure 5.48 Exploded view of optical assembly Insulation baffle plate Wiring terminal bracket M4 hexagon nut Wiring terminal nut Adjusting washer Stop bolt BS-240 Service Manual 5-79 5 Modules and Units 5.9 Photoelectric Unit NOTE: Do not touch the optical surface of the filter with a hand during disassembling/assembling. During disassembling, please wait for at least 5 minutes (if the lamp fan works properly) after turning off the lamp, so as to avoid burn. Do not lose the adjustment washer under the assembly during disassembling. How to do 1 Switch off the power supply of the instrument. 2 Use a cross screwdriver to loosen the screws of the lamp panel, and remove the panel and the panel on the top of the reaction carousel. 3 Refer to the Section about the reaction carousel unit maintenance to remove the upper cover of the reaction chamber and remove the reaction carousel. 4 Remove the right cover. 5 Unplug the power cord of the lamp radiating fan. 6 Unplug the cable between the preamplifier board and the AD collection board. 7 Use a hand to unscrew off the nuts on the cable terminals, and then remove the lamp power cord terminal from the terminals. 8 Use a cross screwdriver to loosen the 2 screws that fasten the wiring terminal bracket, and remove the wiring terminal bracket together with the wiring terminal from the optical assembly. 9 Unscrew the screws that fasten the fiber baffle plate, and remove the fiber baffle plate. 10 Use a hexagon wrench to remove the 3 M4 hexagon screws that fasten the optical assembly. 11 Extract the optical assembly upwards. Adjustment washes may exist under the assembly, which should not be lost and shall be restored after installation. The positions of the washers are shown in the following figure, and the semicircle of the washer clasps the positioning pin. 12 Install the optical assembly in reserved order of the foregoing steps. 5-80 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.9 Photoelectric Unit Figure 5.49 Location of washer Adjusting washer Stop bolt 5.9.5 Replacing Filter When to do Replace the filter when an alarm indicates that the light intensity is low and when energy of a wavelength channel is confirmed weak. Tools Hexagon wrench, cross screwdriver How to do 1 Remove the optical assembly according to the method described in the previous section. 2 According to the following figure and the following table, find the holes of the filter of the corresponding wavelength. Use a cross screwdriver to loosen the retaining screws and washers on both ends of the filter that needs to be replaced, and remove the filter. The outer cylinder surface of the filter has filter center wavelength silkscreen, which may be used to check again whether the filter is expected to be replaced. BS-240 Service Manual 5-81 5 Modules and Units 5.9 Photoelectric Unit Figure 5.50 Exploded view for replacing filter Filter retaining screw Filter retaining washer Filter retaining washer Filter hole No. Hole 1 Filter wavelength (nm) 340 2 3 4 5 6 7 8 405 546 670 450 510 578 630 3 Install a new filter in reversed order of the foregoing steps. 4 Install the optical assembly in reserved order of the foregoing steps. 5 Select "Maintenance"->"Alignment" -> "Photoelectric Gain Adjustment". For details, see the Section about alignment. NOTE: Do not touch the optical surface of the filter with a hand during disassembling/assembling. 5.9.6 Replacing Preamplifier Board When to do The preamplifier board may need to be replaced when no photoelectric signal is output or a dark current is abnormal. Tools Cross screwdriver and hexagon wrench How to do 5-82 1 Remove the optical assembly according to the method. 2 See the following figure. Use a screwdriver to remove the retaining screws of the shielding cover, and remove the shielding cover. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.9 Photoelectric Unit 3 Use a cross screwdriver to remove the retaining screws of the preamplifier board, and gently take out the preamplifier board. 4 Install the preamplifier board in reversed order. Use water as sample or reagent to request test, the reaction curve must be flat and straight. Figure 5.51 Exploded view for installing related parts on preamplifier board side on optical base Optical base Retaining screws on the shielding cover Leaf spring Lens base Lens Washer Preamplifier board Preamplifier board retaining screw Shielding cover NOTE: The parts such as the lens and the lens base in the holes on the preamplifier board in the optical base are fixed by using only the preamplifier board, and are not fixed by other means. Therefore, the preamplifier board should be taken out upwards. When placing the optical assembly after the preamplifier board is taken out, the mounting holes should face upwards to prevent dropping of the parts in the holes. Alignment and confirmation Refer to the Section about alignment. 5.9.7 Replacing lamp fan When to do Replace the lamp radiating fan when it fails. Tools Flathead screwdriver, hexagon wrench BS-240 Service Manual 5-83 5 Modules and Units 5.9 Photoelectric Unit How to do 1 Remove the optical assembly according to the method. 2 See the following figure. Use a screwdriver to remove 4 screws that fasten the fan, and remove the fan. 3 Install a new fan, and fasten it with screws. NOTE: The fan radiates heat towards the lamp base. During installation, the fan surface that bears characters faces the optical base, and the surface without characters faces outwards. Figure 5.52 Exploded view of lamp radiating fan Lamp base Light source radiating fan Fan retaining screws 5.9.8 Replacing Reaction Cuvette When to do Replace a cuvette if, it is detected abnormal through the Cuvette Check procedure; or scratches or cracks are found on the optical surface of the cuvette. WARNING While installing the reaction cuvettes, exercise caution to avoid scratching them. Do not touch the optical surface of the reaction cuvettes. If the optical surface is polluted, the obtained absorbance may be inaccurate. Wear gloves free of fiber and powder to avoid polluting the optical surface of the reaction cuvettes. 5-84 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.9 Photoelectric Unit BIOHAZARD Wear gloves and lab coat, and if necessary, goggles during the maintenance process. Tools Cross screwdriver, fiber-free gloves, dry cloth or gauze, reaction cuvettes, and concentrated wash solution manufactured by our company How to do 1 Select Utility-Maintenance-Maintenance- Biochemistry Maintenance. 2 Choose Replace Cuvette. 3 Select Next. 4 Type in the position number of the cuvette segment you want to replace. The input range is 1-8. Only one position number can be entered each time. 5 Select Replace. 6 Push open the cuvette window cover. Loosen the cuvette anchor plate screw with a hand, flip the anchor plate screw rightward, and take out the reaction cuvette. 7 Install a new reaction cuvette, flip back the reaction cuvette anchor plate, and tighten it with a hand. Note: If all or many segments need to be replaced, the analyzing unit or the mains supply may be turned off first. Push open the cuvette window. Rotate the reaction carousel with a hand to replace the reaction cuvette. 8 After completion of replacing the cuvette, close the cuvette window cover. 9 Select "Utility"->"Maintenance"->"Maintenance"->"Biochemical Maintenance Instruction"->"Reaction Cuvette Detection" to check whether the new cuvette meets the requirements. Note: If the cuvette detection is not performed, the water blank of the new cuvette may be too big to use, or an alarm "Water blank out of range (10X)" appears. 5.9.9 Replacing AD collection board When to do The AD collection board may need to be replaced when no photoelectric signal is output or a dark current is abnormal. Maintenance Tools Cross screwdriver and antistatic gloves How to do 1 BS-240 Service Manual Power off the analyzing unit or the instrument. 5-85 5 Modules and Units 5.9 Photoelectric Unit 2 Use a cross screwdriver to remove the right cover of the instrument. 3 See the following figure. Use a cross screwdriver to remove the screws that fasten the upper cover of the AD box, and remove the upper cover of the AD box. 4 Unplug the signal cable between the AD collection board and the main control board. 5 Use a cross screwdriver to remove the screws that fasten the AD collection board, and remove the AD collection board. 6 Install a new AD collection board in reversed order of the foregoing steps. Figure 5.53 Exploded view of AD collection board Signal cable interface AD collection board AD box upper cover NOTE: Wear ESD gloves when replacing the AD collection board. Alignment and confirmation Refer to the Section about alignment. 5-86 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.10 ISE Unit (Optional) 5.10 ISE Unit (Optional) 5.10.1 Module Functions The ISE module is an optional module for the fully-automated chemistry analyzer and designed to measure the concentration of K+, Na+ and Cl- in serum, plasma and diluted urine. The sample volume for measurement is: 70µl for serum and plasma, 140µl for urine diluted at the ratio of 1:9 (1 for urine and 9 for diluent). 5.10.2 Structure and FRU List The ISE unit consists of the ISE module, pump module and reagent module,functions of which are introduced in the following table. Table 5.10 Components of ISE Unit Name Functions ISE module The ISE module has the following electrodes: Na+, K+, Cl-, and reference. It analyzes and measures sample dispensed through the injection port. Pump module Three peristaltic pumps are contained to transfer reagent and liquid waste. Reagent module The reagent module integrates calibrator A, calibrator B, waste container, and a chip used for monitoring reagent inventory. It provides reagent for sample analysis and stores the liquid waste herein. The figure below shows the composition of the module. BS-240 Service Manual 5-87 5 Modules and Units 5.10 ISE Unit (Optional) Figure 5.54 Exploded view for installation of Measuring module ISE shielding box ISE shielding cover Figure 5.55 Exploded view for installation of pump module Peristaltic Pump 5-88 Pump bracket BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.10 ISE Unit (Optional) Figure 5.56 Exploded view for installation of reagent module Figure 5.57 ISE measuring system tube connection Calibrator B Calibrator A Waste well Pump B Pump W Pump A ISE Table 5.11 List of materials No BS-240 Service Manual FRU No. or Part No. Part Name Remark 1 115-004626-00 K electrode FRU 2 115-004627-00 Na electrode FRU 3 115-004630-00 Cl electrode FRU 4 115-004625-00 Reference electrode FRU 5 115-004628-00 Spacer FRU 6 BA34-10-63657 Main control board of the ISE module FRU 7 BA34-10-63659 Pre-amplification board of the ISE module FRU 8 BA34-10-63658 Compression plate FRU 9 801-BA34-00104-00 ISE module bubble detector FRU 10 801-BA34-00103-00 Sample tube of ISE module FRU 11 BA34-10-63663 O ring of ISE module (3) FRU 12 BA34-10-63668 O ring of ISE electrode FRU 13 082-000684-00 Peristaltic Pump FRU 14 801-BA34-00105-00 The peristaltic pump tube(3 pcs) FRU 15 BA34-10-63666 ISE waste tube connector FRU 5-89 5 Modules and Units 5.10 ISE Unit (Optional) 16 BA34-10-63667 Tube connector of ISE (4) FRU 17 801-BA34-00106-00 ISE module plastic tube for A ,B and W liquid FRU 18 BA34-10-63812 ISE module wand FRU 5.10.3 Unclogging Waste Tubes When to do If the sample contains insoluble substances like fibrin or other substances, they can be accumulated in the waste tube after long time use, causing clogging of the waste tube. Maintenance Tools Name Unclogging tool for ISE Code 115-023372-00 Quantity 1 WARNING Do not spill liquid on the analyzer. Liquid ingression may cause equipment damage. BIOHAZARD Wear gloves and lab coat, and if necessary, goggles during the maintenance process. NOTE Excessive bleach and deionized water flushed into the ISE reagent pack waste bag may cause waste bag over expansion and clog the Cal A & Cal B reagent flow. To prevent this problem, connect to an old used-up reagent pack or use the connector of the used-up reagent. How to do 1 Ensure the analyzer is on idle (standby) condition. Open the ISE cover on the right side analyzer panel. 2 Remove the electrode housing cover. Remove the waste tube fitting from the bottom of the right angle adaptor. Remove waste peri-pump tube from the pump bracket. Refer to pictures below. 3 Connect the waste tube fitting to a syringe and unclogging tool with 5 mL of undiluted household bleach. Refer to pictures below. 4 Press the wand release button to remove the wand from the current in use ISE reagent pack and keep it in a save place. Engage the wand to an old used-up reagent pack. 5-90 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.10 ISE Unit (Optional) 5 Inject bleach into the ISE waste tube and soak the tube for 5 minutes. Discharge the waste into the reagent pack. Note: When the bleach cannot be injected into the ISE pack, remove the wand and push down to open the waste valve manually with a sharp object, and then inject again. If bleach can go through this time, the waste bag was clogged and cannot be used. If bleach still cannot be injected, replacing the ISE wand is recommended. 6 Repeat this step with 5 mL of deionized water without the 5 minutes of soaking time. 7 Remove the wand from the old use-up pack and re-install it back to the current in use ISE pack. Re-install the waste tube fitting back to the ISE electrode housing right angle adaptor and waste peri-pump tube back to the pump bracket. Re-install the housing cover. Alignment and confirmation Perform ISE pump calibration and if it succeeds, the tube has been successfully unclogged. 5.10.4 Replacing Pump Tube When to do Replace the pump tube when it is aged or leaking or has not been maintained for a half year. Maintenance Tools Cross screwdriver, pump tube. Precautions: BIOHAZARD Wear gloves and lab coat, and if necessary, goggles during the maintenance process. How to do 1 Ensure the analyzer is on idle (standby) condition. Open the ISE cover on the right side analyzer panel. 2 Execute Utility-Maintenance-Maintenance- ISE Maintenance – Reagent Unpack. 3 Execute Utility-Maintenance-Maintenance- ISE Maintenance – Replace Pump Tube and Calibrator Tube. 4 Select "Pump Tube". 5 Replace it with new one. Make sure the connectors and tubes are correctly connected; otherwise ISE module failure may occur. 6 Load the reagent pack and the procedure is completed. 7 Install back the right side panel. BS-240 Service Manual 5-91 5 Modules and Units 5.10 ISE Unit (Optional) 8 Record the maintenance log. Alignment and confirmation Perform ISE pump calibration and if it succeeds, the tube has been successfully replaced. 5.10.5 Replacing Calibrator Tube When to do Replace the calibrator tube when it is aged or leaking or has not been maintained for over one year. Maintenance Tools Cross screwdriver, calibrator tube Precautions: BIOHAZARD Wear gloves and lab coat, and if necessary, goggles during the maintenance process. How to do 1 Ensure the analyzer is on idle (standby) condition. Open the ISE cover on the right side analyzer panel. 2 Execute Utility-Maintenance-Maintenance- ISE Maintenance – Reagent Unpack. 3 Execute Utility-Maintenance-Maintenance- ISE Maintenance – Replace Pump Tube and Calibrator Tube. 4 Select Calibrator Tube. 5 Replace it with new one. Make sure the connectors and tubes are correctly connected; otherwise ISE module failure may occur. 6 Load the reagent pack and the procedure is completed. 7 Install back the right side panel. 8 Record the maintenance log. Alignment and confirmation Perform ISE pump calibration and if it succeeds, the tube has been successfully replaced. 5-92 BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.10 ISE Unit (Optional) 5.10.6 Inventory Calculation of Reagent Pack Reagent pack as one part of the ISE module should be installed as instructed by this manual. You should install reagent pack while the analyzer is powered off, and then check the reagent inventory and prime the ISE module with Calibrator A and B. If the reagent pack is not properly installed, error may be produced in inventory calculation, and the actual volume may differ from that displayed on the screen. The following operations may result in the actual inventory of the ISE reagent pack less than the displayed one. 1 When the analyzer is turned on and the ISE module is in failure status, the reagent pack is changed, without recovering the module failure or checking the inventory immediately. Operation explanation: When the ISE module is in Stopped status, the software will stop writing volume information into the reagent pack chip. However, the analyzer will automatically execute the sip action of the ISE module during startup to consume certain amount of reagent. Therefore, when the reagent pack is replaced, but the ISE module is not restored or the screen display refreshed, the software will not calculate the reagent consumption volume until reading the new reagent pack. At this moment, the actual reagent volume = displayed volume - consumption of Sip after replacing the reagent pack, which means that the actual reagent volume is less than the displayed volume on the screen. 2 When the analyzer is powered on but the software is closed or the computer is shut down, the reagent pack is changed. Operation explanation: During shutdown, the analyzer will automatically execute the sip action of the ISE module to consume certain amount of reagent. When the analyzer is started up, the software will detect a new reagent pack and zero the consumption that is not written into the reagent pack chip, and then display the volume read from the reagent pack chip. At this moment, the actual reagent volume = displayed volume - consumption of Sip after replacing the reagent pack, which means that the actual reagent volume is less than the displayed volume on the screen. 3 When the analyzer is powered on but the software is closed or the computer is shut down, the reagent pack is switched to another analyzer and then installed back after use. Operation explanation: During shutdown process, the analyzer cannot detect if the reagent pack is normal. If the reagent pack is installed on the current instrument after being used on another one, the analyzer, when started up, will detect that the recorded reagent volume on the chip is less than the original one, and the software will zero the reagent consumption of Sip during shutdown and display the read volume on the screen. At this moment, the actual reagent volume = displayed volume - consumption of Sip during shutdown (with reagent pack installed), which means that the actual reagent volume is less than the displayed volume on the screen. BS-240 Service Manual 5-93 5 Modules and Units 5.10 ISE Unit (Optional) 4 The ISE module is de-configured, but the reagent pack is still on the instrument. Operation explanation: Though the ISE module is de-configured, the original reagent pack information is not cleared from the database. When the ISE module is configured again with the same reagent pack, the Sip period without ISE module will be included in calculation of the reagent consumption. NOTE: In condition 1, restore the ISE module and check the reagent inventory immediately to minimize or eliminate the calculation error. In condition 2 and 3, before running the operating software to start up the analyzer, remove the reagent pack and then run the operating software. After the startup procedure is complete, re-install the reagent pack and inquire the reagent volume immediatelyto minimize or eliminate the calculation error. In condition 4, remove the reagent pack and then de-configure the ISE module. The following operations may result in the actual inventory of the ISE reagent pack more than the displayed one. 1 When the analyzer is turned on and the ISE reagent pack is changed, without checking the inventory immediately. Operation explanation: The analyzer will automatically refresh the ISE reagent consumption, by subtracting the consumption of calibrator A or B from the known volume (of the original reagent pack) and writing it into the reagent pack chip, when calibrator A or B is consumed for 1%. The written volume is based on the original reagent pack before replacement, and the volume of a used reagent pack is often less than that of a new one. So the volume recorded on the chip of the new reagent pack is less than the actual volume, and the actual volume is greater than the displayed volume on the screen. 2 When the analyzer is powered on but the software is closed or the computer is shut down, the reagent pack is removed for storage, and re-installed before the operating software is run. Operation explanation: During shutdown process, the analyzer cannot detect if the reagent pack is normal. If the reagent pack is loaded before the analyzer is started up, the analyzer, when started up, will detect that the reagent volume recorded on the chip is same as that recorded by the software, but the software will confirm by mistake the reagent consumption of sip during shutdown. (Actually, no reagent is consumed.)Therefore, after calculating the reagent consumption, the software will write the new volume information onto the chip and display it on the screen. At this moment, the actual reagent volume = displayed volume + consumption of Sip during shutdown (with reagent pack installed), which means that the actual reagent volume is greater than the displayed volume on the screen. 3 5-94 The analyzer is powered off after the operating software is closed, and it is powered on again before the operating software is run. BS-240 Chemistry Analyzer Service Manual 5 Modules and Units 5.10 ISE Unit (Optional) Operation explanation: When closed normally, the operating software cannot detect if the ISE module is working. When powered on before the software is run, the analyzer, when started up, will detect that the reagent volume recorded on the chip is same as that recorded by the software, but the software will confirm by mistake the reagent consumption of sip during shutdown. (Actually, no reagent is consumed because the ISE module is not working while the analyzer is powered off.)Therefore, after calculating the reagent consumption, the software will write the new volume information onto the chip and display it on the screen. At this moment, the actual reagent volume = displayed volume + consumption of Sip during shutdown (with reagent pack installed), which means that the actual reagent volume is greater than the displayed volume on the screen. NOTE: In condition 1, inquire the ISE reagent volume immediatelyto minimize or eliminate the calculation error. In condition 2 and 3, before running the operating software to start up the analyzer, remove the reagent pack and then run the operating software. After the startup procedure is complete, re-install the reagent pack and inquire the reagent volume immediatelyto minimize or eliminate the calculation error. BS-240 Service Manual 5-95 5 Modules and Units 5.10 ISE Unit (Optional) 5-96 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits BS-240 Chemistry Analyzer Service Manual 6-1 6 Hardware Circuits 6.1 Overview 6.1 Overview This chapter describes the functions of the PCBAs (printed circuit board assemblies) used on the BS-240 Chemistry Analyzer. 6-2 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.2 Hazards 6.2 Hazards NOTE: While the instrument is working, do not touch the hardware circuit boards with your hands or other objects. Before removing a circuit board, disconnect the instrument from the (AC) power supply. Please wear a pair of anti-static gloves or take other measures to prevent static electricity prior to removing a circuit board. BS-240 Service Manual 6-3 6 Hardware Circuits 6.3 Summary of PCBAs 6.3 Summary of PCBAs Sequence number and function description of PCBAs The table bellows provides a summary of the PCBAs used on the BS-240 Chemistry Analyzer and briefly describes their functions. Table 6.1 Summary of PCBAs PCBA (PCB) Functions No. Main Control Board 051-002414-00 The main control board is the control center of the BS240. It is mainly used to fulfill the following tasks: communicating with a computer through the RS232 serial port to transmit data and instructions; communicating with the smart modules, including the ISE module, through the extended serial ports to transmit data and instructions; and controlling digital potentiometer adjustment and photoelectric data collection of the AD collection board and receiving the photoelectric data. #1 Power drive board 051-002413-00 Controlling and driving the sample/reagent probe, sample/reagent mixer, sample/reagent carousel, reaction carousel, cuvette wash station, wash syringe, and related parts. #2 Reagent refrigeration board BA20-30-75227 The reagent refrigeration board is used to drive the reagent refrigeration circuit (including heatsinks), fans, and demisting circuit of the reagent bar code reader. #3 Level sense board 051-002479-00 The two boards respectively sense the level of samples and reagents, and detect/convert the signals of the two vertical obstruct detection sensors. #4 AD collection board BA10-30-77759 The AD collection board adjusts the photoelectric signals output by the pre-amplification board and controls it for AD conversion. It provides an SPI interface for connecting with the main control board. #5 Pre-amplification Board BA10-30-77760 The pre-amplification board converts the signals of the discrete photodiode array from analog to digital. #6 Power supply board 051-002404-00 These boards are used to power the whole unit, outputting the A12V, A5V digital, D12V/E12V analog, C12V power, B24V power, and the B12V power for partial Peltiers. #7 For locations of boards in the instrument, see the following exploded view. 6-4 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.3 Summary of PCBAs Figure 6.1 Locations of boards in the instrument (1) Power drive board Main Control Board Power supply board BS-240 Service Manual 6-5 6 Hardware Circuits 6.3 Summary of PCBAs Figure 6.2 Locations of boards in the instrument (2) Level sense board Reagent refrigeration board Pre-amplification Board 6-6 AD collection board BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA 6.4 Functions of PCBA 6.4.1 Control Structure The BS-240 Chemistry Analyzer consists of the analyzing unit (analyzer), operation unit (computer), and output unit (printer). The analyzing unit (analyzer) is composed of the temperature control system, reaction system (including ISE module), photometric system, sample/reagent handling system, mixer system, and cuvette wash station. The general control structure of the BS-240 is shown in the figure below. The hardware system has the following functions: Communicating with a computer through the serial port, and receiving/sending commands, responses, and data. Controlling data collection of the photometric system. Controlling movement and status signal collection of execution units. Controlling working and temperature control signal collection of the temperature control system. Figure 6.3 Overall functions Optical System Photoelectric signal Control signal Temperature control signal PC RS232 Hardware System Temperature control system Control signal Control signal Position signal Drive and execute system 6.4.2 Main Control Board Functions and Principles The main control board (051-002414-00) is used to fulfill the following functions: Communicating with a computer through the serial port to transmit data and instructions and to update the application programs of the PCBA. BS-240 Service Manual 6-7 6 Hardware Circuits 6.4 Functions of PCBA Communicating with the smart modules, including the ISE module, through the extended serial ports to transmit data and instructions. Controlling digital potentiometer adjustment and photoelectric data collection of the AD collection board and receiving the photoelectric data. Detecting position signal and horizontal obstruct signal. Detecting level sense signal and vertical obstruct signal through the interface for the level sense board. The figure below shows the relation between the main control board and other PCBAs. Figure 6.4 Relation between main control board and other PCBAs Power supply board Reagent refrigeration board Main Control Board Power drive board AD collection board Preamplifier board The functional diagram of the main control board is shown below. 6-8 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Figure 6.5 Functional diagram of main control board Bus TTL Serial Port Location sensor Drive clock/direction signal Injection unit Probe wash pump valve signal Bus Level floater Drive clock/direction signal Mixer Unit TTL Serial Port Wash pump valve signal Bus Reaction Unit PC Drive clock/direction signal TTL Serial Port FPGA Drive clock/direction signal Bus Cuvette Wash Unit TTL Serial Port Bus TTL Serial Port Bus Wash pump valve signal Temperature Control Unit Reaction carousel heater control signal Control signal of deionized water heater Control signal of wash solution heater Main Control Unit Main Control Board Description PCB layout The PCB layout of the main control board is shown below. BS-240 Service Manual 6-9 6 Hardware Circuits 6.4 Functions of PCBA Figure 6.6 PCB layout of main control board Connectors The main control board includes the following connectors. Power supply: Power supply input (J2): 6-pin, providing +12V analog, +5V digital for the PCBAs. Pin No. Signal Reference Value 1 +5V 4.75~5.25V 2 GND / 3 / / 4 / / 5 12V 11.4~-12.6V 6 GND / Connectors for sending/receiving communication signals: 6-10 Connector (J7) for PC: 10-pin, RS232, used for communication with a computer. Connector (J1) for drive signal: 40-pin, TTL, used for transmitting a control signal to the power drive board. Connector (J39) for AD collection board: DB25, used for communication with the AD collection board and providing power supply for it. Connector (J6) for ISE module: 8-pin, RS232, used for communication with the ISE module. BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Connector (J5) for bar code module: 8-pin, RS232, used for communication with the bar code module. Connector (J12) for photocoupler: 36-pin, used for input of movement station position sensor. Connector (J13) for photocoupler: 36-pin, used for input of movement station position sensor. Connector (J14) for photocoupler: 10-pin, used for input of level floater sensor. Connector (J18) for temperature sensor: 4-pin, used for input of reaction carousel temperature sensor. Connector (J21) for temperature sensor: 4-pin, used for input of deionized water temperature sensor. Connector (J22) for temperature sensor: 4-pin, used for input of wash solution temperature sensor. Connector (J23) for temperature sensor: 4-pin, used for input of reagent preheating temperature sensor. Connector (J11) for power supply and signal output of level sense board: 4-pin, used to input a level signal and an anti-collision signal into and provide a power supply for the level sense board. Connectors for debugging: JTAG connector (J8): 10-pin, used for debugging the FPGA. Indicators The main control board contains the following indicators. +12V power supply indicator (D8): green. It is lit when the analyzer power switch is turned on, indicating that the +12V power supply has been connected. +5V power supply indicator (D7): green. It is lit when the analyzer power switch is turned on, indicating that the +5V power supply has been connected. +3.3V power supply indicator (D9): green. It is lit when the analyzer power switch is turned on, indicating that the +3.3V power supply has been connected. Analog +12V power supply indicator (D20): green. It is lit when the analyzer power switch is turned on, indicating that the analog +12V power supply has been connected. Analog -12V power supply indicator (D21): green. It is lit when the analyzer power switch is turned on, indicating that the analog -12V power supply has been connected. D19: orange FPGA configuration indicator. On indicates FPGA is successfully configured. D14: Green FPGA working status indicator. LED is flashing every second, indicating FPGA works normally. BS-240 Service Manual 6-11 6 Hardware Circuits 6.4 Functions of PCBA D23: Green. When communication with the lower layer unit is interrupted, the indicator is lit. Test points In the following positions of the main control board can signal tests be performed. 12V: +12V power supply input. Normal range: 11.4 - 12.6V. P12V: analog +12V power supply input. Normal range: 11.4 - 12.6V. N12V: analog -12V power supply input. Normal range: -11.4 - -12.6V. 5V: +5V power supply input. Normal range: 4.75 - 5.25V. VDD: +3.3V power supply. It is secondary power supply used to power the major digital circuits of the PCBA. Normal range: 2.97 - 3.63V. 2.5V: +2.5V power supply. It is secondary power supply used for FPGA downloading. Normal range: 2.25 - 2.75V. 1.2V: +1.2V power supply. It is secondary power supply used to power the FPGA core. Normal range: 1.14 - 1.26V. Installation Methods and Precautions NOTE: Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static gloves or wrist straps. Make sure that the connectors are inserted properly into the PCBA. Check the connectors with locks and ensure they have been locked properly. Check other connectors and ensure that they are inserted into the end of the slots. It requires relatively great force to plug/unplug connectors J1 and J12~J13. Hold the PCBA by its edge while plugging/unplugging these connectors to prevent it from being deformed or damaged. After connecting J39 connector (DB25), tighten the retaining screws on two sides of it. Parameters should be saved before uninstalling the board, and can be loaded after the board is replaced. 6.4.3 Power drive board Description 6-12 The power drive board controls and drives the step motors, DC motors and valves. Receiving action signals from the main control board through the relevant interface. BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Controlling and driving movements of the wash station, wash syringe, sample/reagent/reagent carousel, sample/reagent/reagent probe, mixer, reaction carousel, valves, etc. The functional diagram of the power drive board is shown below. Figure 6.7 Power drive board Power drive board Bus TTL Serial Port Location sensor Injection unit Drive clock/direction signal Probe vertical motor Probe wash pump valve signal Probe horizontal motor Probe syringe motor Bus Level floater Mixer Unit TTL Serial Port Drive clock/direction signal Wash pump valve signal CPLD Motor drive A3985/A3 979 Bus Reaction Unit PC Drive clock/direction signal Wash syringe motor TTL Serial Port FPGA Drive clock/direction signal Bus Cuvette Wash Unit TTL Serial Port Bus TTL Serial Port Bus Wash pump valve signal Pump/ Valve Drive Pump/ valve Pump/Valve Drive Temperature Control Unit Reaction carousel heater control signal Control signal of deionized water heater Control signal of wash solution heater Main Control Unit 220V heater drive Reaction Carousel Heater 220V heater drive Deionized Water Heater 220V heater drive Wash Solution Heater Power supply board Main Control Board Description The PCB layout of the power drive board is shown below. BS-240 Service Manual 6-13 6 Hardware Circuits 6.4 Functions of PCBA Figure 6.8 Power drive board PCB Connectors: The power drive board contains the following connectors. Power supply: Power supply input (J17): 8-pin, providing analog 24V and 12V and digital +5V for the PCBAs. Pin No. Signal Reference Value 1 +24V 23.5~26V 2 +24V 23.5~26V 3 +5V 4.75~5.25V 4 +12V 11.4~12.6V 5 GND / 6 GND / 7 GND / 8 GND / Connectors for sending/receiving communication signals: Connector (J20) for transmitting control signal of main control board: 40-pin, TTL, used for transmitting a control signal. Connectors for debugging: JTAG connector (J2): 10-pin, used for debugging the FPGA. Drive control connectors: 6-14 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Valve driving connector (J6): 8-pin, used for driving interior pump and interior wash valve of sample/reagent probe. Valve driving connector (J9): 8-pin, used for driving phase-1\2\3\4 waste pump, deionized water and wash solution dispense valves, and mixer exterior valve. Sample/reagent carousel motor driving connector (J15): 4-pin, used for driving the sample/reagent carousel motor. Reaction carousel motor driving connector (J16): 4-pin, used to drive reaction carousel motor. Sample/reagent horizontal motor driving connector (J5): 4-pin, used for driving the sample/reagent horizontal motor. Sample/reagent vertical motor driving connector (J13): 4-pin, used for driving the sample/reagent vertical motor. Sample/reagent syringe motor driving connector (J8): 4-pin, used for driving the sample/reagent syringe motor. Deionized water syringe motor driving connector (J11): 4-pin, used for driving the deionized water syringe motor. Wash solution syringe motor driving connector (J12): 4-pin, used for driving the wash solution syringe motor. Wash station motor driving connector (J4): 4-pin, used for driving the wash station motor. Sample/reagent mixer horizontal motor driving connector (J10): 4-pin, used for driving the sample/reagent mixer horizontal motor. Sample/reagent mixer vertical motor driving connector (J14): 4-pin, used for driving the sample/reagent mixer vertical motor. Mixer motor driving connector (J22): 2-pin, used to drive the mixer motor. Reagent preheating driving connector (J24): 2-pin, used for driving the reagent preheater. Indicators The power drive board contains the following indicators. Power supply: +12V power supply indicator (D3): green. It is lit when the analyzer power switch is turned on, indicating that the +12V power supply has been connected. +5V power supply indicator (D2): green. It is lit when the analyzer power switch is turned on, indicating that the +5V power supply has been connected. +24V power supply indicator (D1): green. It is lit when the analyzer power switch is turned on, indicating that the +24V power supply has been connected. Test points BS-240 Service Manual 6-15 6 Hardware Circuits 6.4 Functions of PCBA In the following positions of the power drive board can signal tests be performed. 12V: +12V power supply input. Normal range: 11.4 - 12.6V. 24V: 24V power supply input. Normal range: 23.5 - 26V. 5V: +5V power supply input. Normal range: 4.75 - 5.25V. Installation Methods and Precautions NOTE: Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static gloves or wrist straps. Make sure that the connectors are inserted properly into the PCBA. Check the connectors with locks and ensure they have been locked properly. Check other connectors and ensure that they are inserted into the end of the slots. It requires relatively great force to plug/unplug connectors J20 and J17. Hold the PCBA by its edge while plugging/unplugging these connectors to prevent it from being deformed or damaged. 6.4.4 AD collection board Functions and Principles The AD collection board is intended to implement the following functions: Adjusting gains of 12 photoelectric signals via a digital potentiometer. Controlling the channel selection switch to switch among 12 channel signals at different time. Converting photoelectric analog signals into digital signals via an AD converter and then outputting the signals. The figure below shows the relation between the AD collection board and other PCBAs. Figure 6.9 Relation between AD collection board and other PCBAs Preamplifier board AD collection board Main Control Board 6-16 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA The functional diagram of the AD collection board is shown below. Figure 6.10 Functional diagram of AD collection board Preamplifier board Single analog signal gain adjustment Main Control Board ADC Description PCB layout The PCB layout of the AD collection board is shown below. Figure 6.11 PCB layout of AD collection board Connectors The AD collection board contains the following connectors. Power supply: BS-240 Service Manual 6-17 6 Hardware Circuits 6.4 Functions of PCBA Power supply output connector (J1): 6-pin, used to provide +12V and -12V power supplies for the preamplifier board and receive optical and electric signals from the preamplifier board. Pin No. Signal Reference Value 1 Signal 0.25V-2.5V (water blank condition) 2 +12V 11.4~12.6V 3 GND / 4 GND / 5 GND / 6 -12V -11.4~-12.6V Main control board connector (P1): 25-pin, used for communication with the main control board. Indicators The AD collection board contains the following indicators. Power supply: +12V power supply indicator (D5): green. When it is lit, it indicates that the +12V power supply has been connected. -12V power supply indicator (D4): green. When it is lit, it indicates that the -12V power supply has been connected. +5V power supply indicator (D3): green. When it is lit, it indicates that the +5V power supply has been connected. Test points In the following positions of the AD collection board can signal tests be performed. VIN: Optical/electric signal output Normal range: varies with the signal strength and lies between 0-5V. VPP: +12V power supply input. Normal range: 12V±5%, that is, 11.4 - 12.6V. VSS: -12V power supply input. Normal range: -12V±5%, that is, -11.4 - -12.6V. VCC: +5V power supply input. Normal range: 5V±5%, that is, 4.75 - 5.25V. AVCC: analog +5V power supply. Normal range: 5V±5%, that is, 4.75 - 5.25V. DVCC: digital +5V power supply. Normal range: 5V±5%, that is, 4.75 - 5.25V. Installation Methods and Precautions NOTE: Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static gloves or wrist straps. 6-18 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Make sure that the connectors are inserted properly into the PCBA. Check the connectors with locks and ensure they have been locked properly. Check other connectors and ensure that they are inserted into the end of the slots. It requires relatively great force to plug/unplug connectors J1. Hold the PCBA by its edge while plugging/unplugging these connectors to prevent it from being deformed or damaged. After connecting P1 connector (DB25), tighten the retaining screws on two sides of it. 6.4.5 Pre-amplification Board Functions and Principles The pre-amplification board is intended to implement the following functions: Converting optical signals into electric signals via a photodiode array. With the help of the back circuit, filtering signals and transmitting them to the AD collection board after amplification. The figure below shows the relation between the pre-amplification board and other PCBAs. Figure 6.12 Relation between pre-amplification board and other PCBAs Optical signal Preamplifier board ADC collection board The functional diagram of the pre-amplification board is shown below. Figure 6.13 Functional diagram of preamplifier board Optical signal PDA Signal filtering and amplification ADC collection board BS-240 Service Manual 6-19 6 Hardware Circuits 6.4 Functions of PCBA Description PCB layout The preamplifier board layout of the main control board is shown below. Figure 6.14 PCB layout of preamplifier board Connectors The pre-amplification board includes the following connectors. Power supply: Power supply input connector (2): 1-pin, used to provide +12V power supplies for the preamplifier board. Power supply input connector (6): 1-pin, used to provide -12V power supplies for the preamplifier board. Power supply input connector (3): 1-pin, used to ground the preamplifier board. Connectors for sending/receiving communication signals: 6-20 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Preamplifier board signal output connector (1): 1-pin, used to transmit 1 optical/electric analog signal. Installation Methods and Precautions Since encapsulated inside the Optical measurement assembly, the preamplifier board should be maintained together with the Optical measurement assembly rather than maintained independently. 6.4.6 Level sense board Functions and Principles The BS-240 level sense board is used to sense the sample/reagent fluid level, and primarily implements the following functions: Sensing the reagent/sample levels; detecting the fluid level steadily and reliably, especially allowing the sample/reagent probe to correctly detecting the fluid level inside reaction cuvettes. Outputting level sense signals to the main control board when the probe contacts the fluid level. Detecting vertical collision, and outputting the detection signal to the main control board. The functional diagram of the level sense board is shown below. Figure 6.15 Functional diagram of level sense board Vertical collision circuit Dual tube probe Capacity detection chip CPU Connect or circuit Main Control Board Description PCB layout The top layer of the level sense board is shown below. BS-240 Service Manual 6-21 6 Hardware Circuits 6.4 Functions of PCBA Figure 6.16 Level detection board PCB Connectors The level sense board contains the following connectors. Probe connector (J1): 2-pin, used for connecting the sample/reagent probes with related circuit. Pin No. Signal 1 GND 2 Probe capacitor signal input Power supply and signal output connector (J2): 4-pin, used to provide power supply for the level sense board and output the level sense signal and vertical collision signal. Pin No. Signal 1 GND 2 Vertical collision output Reference Value / signal High level (about 4V) is output when no vertical collision occurs, and low level (about 0V) is output when vertical collision happens. 3 Level sense signal output Low level (about 0V) is output when the probe fails to detect the fluid level, and high level (about 4V) is output when the probe detects the fluid level. 4 +12V 11.4~12.6V Serial port communication cable connector (J3): 3-pin, used for communication between the level sense board and the control drive board. Pin No. Signal 1 RXD 2 RST 3 TXD Indicators The level sense board contains the following indicators. 6-22 Level sense indicator (D5): green. It is extinguished when the probe fails to detect the fluid level, and vice versa. Vertical collision indicator (D1): green. It is extinguished when no vertical collision occurs, and vice versa. BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Level sense board operation indicator (D2): yellow. The level sense board operation indicator is blinking when the board works normally, and is always lit when strong interference is detected. Test points In the following positions of the level sense board can signal tests be performed. LEVEL: level sense signal output. Normal condition: Low level (about 0V) is output when the probe fails to detect the fluid level, and high level (about 4V) is output when the probe detects the fluid level. GND: grounding terminal of the level sense board. 5V: 5V power supply of the level sense board, with a normal range of 4.75 - 5.25V. 3.3V: 3.3V power supply of the level sense board, with a normal range of 2.97 - 3.63V. RAM: vertical collision signal output. Normal condition: High level (about 4V) is output when no vertical collision occurs, and low level (about 0V) is output when vertical collision happens. Installation Methods and Precautions NOTE: Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static gloves or wrist straps. Make sure that the connectors are inserted properly into the end of the slots on the PCBA. It requires great force to plug/unplug the connectors. Hold the PCBA by its edge while plugging/unplugging the connectors to prevent it from being deformed or damaged. 6.4.7 Reagent Refrigeration Board Functions and Principles The reagent refrigeration board is relatively independent in the entire hardware system and has the following major functions: Refrigeration control and drive: controlling the switch of the heatsinks and refrigerating the reagent compartment to keep the temperature within certain range. Control of cooling fans and demisting. The functional diagram of the reagent refrigeration board is shown below. BS-240 Service Manual 6-23 6 Hardware Circuits 6.4 Functions of PCBA Figure 6.17 Functional diagram of reagent refrigeration board Temperature sensor Reagent refrigeration board Temperature indication circuit Refrigeration control circuit 2 Peltiers 2 Peltier radiating fans Voltage conversion circuit 12 V 6 reserved system radiating fans 1 lamp radiating fan 12VFAN 1 power drive board radiating fan Antifogging circuit Main Control Board Photoelectric coupling Reaction carousel fan feedback signal processing circuit 1 reserved radiating fan 4 reaction carousel temperature control fans Description PCB layout The PCB layout of the reagent refrigeration board is shown below. Figure 6.18 Reagent refrigeration board PCB Connectors The reagent refrigeration board includes the following connectors. Power supply: 6-24 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA Power supply input connectors (J86): providing refrigeration F12V and refrigeration B12V for the PCBAs. Pin No. Signal Reference Value 1 F12V 11.4~12.6V 4 GND 0 2, 3 B12V 11.4~12.6V 5, 6 FGND / Description 12V power supply input for cooling fans. Refrigeration power supply input: used for loading the corresponding Peltiers J89_1 and J89_2. Peltier connectors (J4-J7): providing power supply for the Peltiers. Connector Pin No. J89_1 1 Signal Reference Value 1 Peltier 0V for ON and 12V for OFF. 2 Connector J89_2 12V Pin No. 1 Signal Reference Value 2 Peltier 0V for ON and 12V for OFF. 2 12V Demisting circuit connector (J15): providing power supply for the demisting heater. Connector J9 Pin No. 1 Signal 1 demisting heaters 2 Reference Value 11.4~12.6V / 3 / 4 0 Reagent refrigeration fan connectors (J84, J85): providing a power supply for the reagent refrigeration 12V fan. Connector J84 J85 Pin No. Signal Reference Value 1 GND / 2 FAN 11.4~12.6V 1 GND / 2 FAN 11.4~12.6V Temperature sensor connector (J88): checking temperature of the refrigeration compartment and refrigeration board. BS-240 Service Manual 6-25 6 Hardware Circuits 6.4 Functions of PCBA Connector Pin No. J88 Signal Reference Value 1 GND 0V 2 Temperature sensor signal of the refrigeration compartment / PCBA radiating fan connector (J75): providing a 12V power supply for the PCBA radiating fan. Connector J75 Pin No. Signal Reference Value 1 GND / 2 FAN 11.4~12.6V Whole unit radiating fan connector (J81): providing a 12V power supply for the whole unit radiating fan. Connector J81 Pin No. Signal Reference Value 1 GND / 2 FAN 11.4~12.6V Lamp radiating fan connector (J73): providing a 12V power supply for the lamp radiating fan. Connector J73 Pin No. Signal Reference Value 1 GND / 2 FAN 11.4~12.6V Reaction carousel constant-temperature fan connectors (J76, J77): providing a 12V power supply for the fan that keeps temperature of the reaction carousel. Connector J76 J77 Pin No. Signal Reference Value 1 GND / 2 FAN 11.4~12.6V 1 GND / 2 FAN 11.4~12.6V Indicators The reagent refrigeration board includes the following indicators. Power supply: 6-26 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.4 Functions of PCBA No. Description Color Introduction D45 C24V Green When it is lit, it indicates that the power supply works normally. D48 VCC Green When it is lit, it indicates that the power supply works normally. D47 B12V Green When it is lit, it indicates that the power supply works normally. D49 VDD Green When it is lit, it indicates that the power supply works normally. D46 F12V Green When it is lit, it indicates that the power supply works normally. Indicators to indicate PCBA functions: No. Description Color Introduction D15 Fan power indicator supply Green When it is lit, it indicates that the refrigeration board is working. D3 Refrigeration power supply indicator Green When it is lit, it indicates that the refrigeration board is working. D4 Indicating the reagent compartment temperature Red When the reagent compartment temperature exceeds the high limit D6 Indicating the reagent compartment temperature Green When the reagent compartment temperature is within the target range D5 Indicating the reagent compartment temperature Yellow When the reagent compartment temperature is lower than the low limit D7 Indicating status the peltier Green When it is lit, it indicates that the peltier is working. of Installation Methods and Precautions Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static gloves or wrist straps. Make sure that the connectors are inserted properly into the PCBA. Check the connectors with locks and ensure they have been locked properly. Check other connectors and ensure that they are inserted into the end of the slots. Installation Methods and Precautions NOTE: Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static gloves or wrist straps. Make sure that the connectors are inserted properly into the end of the slots on the PCBA. BS-240 Service Manual 6-27 6 Hardware Circuits 6.4 Functions of PCBA It requires great force to plug/unplug the connectors. Hold the PCBA by its edge while plugging/unplugging the connectors to prevent it from being deformed or damaged. 6-28 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.5 Power Supply System 6.5 Power Supply System 6.5.1 Power Supply System of Whole Unit Figure 6.19 Power supply system of the whole unit Connected to housing E yellow green L brown Power jack Analyzer Switch Main Power Switch L brown N light blue L N light blue N Power supply board Power supply board The power supply board, as the major part of the entire power supply system, provides power supply for the lamp, DC power supply for major control boards, and power supply for refrigeration modules. For details, see the figure below: Figure 6.20 Power supply board A12V/2A Lamp Power supply board 5V Main Control Board 5V/4A Analog +/-12V Analog Power Supply Conversion Board Analog +/-12V AD collection board Preamplifier board 12V/5A Power drive board 24V/6A Power Supply Conversion Board 12V/11A ISE module Reagent refrigeration board 6.5.2 Performance Indices Power supply input AC voltage input: 220-240V, 220/230V, and 110/115V. AC voltage frequency: 50/60±1Hz AC input power: 1KVA Power supply output No 1 Name 5V BS-240 Service Manual Rated Output 5V/4A Usage 5V digital Control Description Controlled by analyzer power switch 6-29 6 Hardware Circuits 6.5 Power Supply System 2 12V 12V/2A Lamp Controlled by operating software 3 12V 12V/11A Reagent refrigeration Controlled by power switch 4 12V 12V/5A 12V drive Controlled by analyzer power switch 5 24V 24V/6A 24V drive heater and main Controlled by analyzer power switch NOTE: Control by operating software means that the lamp voltage is controlled by the operating software while the main power switch and analyzer power switch are turned on. Control by analyzer power switch means that the power supply is controlled by the analyzer power switch with the prerequisite that the main power switch is turned on. Control by main power switch means that the power supply works normally when the analyzer power switch is turned on, and vice versa. 6.5.3 Power supply board Functions and principles are described below. The PCB layout of the power supply board is shown below. 6-30 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.5 Power Supply System Figure 6.21 PCB layout of power supply board Indication of voltage outputs: LED BS-240 Service Manual Indication Status Controlled By L12V When L12V is lit, it indicates that the lamp outputs a 12V power. Controlled operating software by D5V When D5V is lit, it indicates that the PCBA outputs a 5V power. Controlled by analyzer power switch B12V When B12V is lit, it indicates that the reagent refrigerator outputs a 12V power. Controlled by power switch P12V When P12V is lit, it indicates that the lamp outputs a 12V power. Controlled by analyzer power switch P24V When P24V is lit, it indicates that the PCBA outputs a 24V power. Controlled by analyzer power switch main 6-31 6 Hardware Circuits 6.5 Power Supply System NOTE: Control by operating software means that the lamp is controlled by the operating software while the main power switch and analyzer power switch are turned on. Control by analyzer power switch means that the power supply is controlled by the analyzer power switch with the prerequisite that the main power switch is turned on. Control by main power switch means that the power supply works normally when the analyzer power switch is turned on, and vice versa. Output connectors: No. J1 J2 J3 Pins Output Description 1 AC input L 3 AC input N 1 12VOUT 2 Analyzer Switch 3 Ground (Analyzer Switch) 1 deionized water Heater (220V) 12 2 Wash solution heater (220V) 5 3 10 7 9 J4 6-32 Upper heater (220V) of reaction carousel Lower heater (220V) of reaction carousel 4 / 6 / 8 / 11 / 1 Lamp ground 2 Ground 3 Ground 4 Ground 5 Ground 6 Ground 7 Ground 8 Ground 9 Ground 10 Ground BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.5 Power Supply System No. J5 J6 Pins Output Description 11 12V output of lamp 12 12V output of reagent refrigeration 13 12V output of reagent refrigeration 14 24V output of driving 15 24V output of driving 16 12V output of driving 17 12V output of driving 18 12V output of driving 19 5V output 20 5V output 1 Control input of deionized water heater 2 Control input of wash solution heater 3 Control heater 4 Ground 5 Control input of inlet valve 6 Ground 7 Control input of SV07 dispense valve 8 Ground 9 Reserved 10 Ground 11 Lamp control input 12 Ground 1 Driving output of inlet valve input of reaction carousel 2 3 Driving output of SV07 dispense valve 4 5 Reserved 6 J7 1 Radiating fan 2 3 Radiating fan 4 BS-240 Service Manual 6-33 6 Hardware Circuits 6.5 Power Supply System 6-34 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.6 Wiring Diagram of Analyzer 6.6 Wiring Diagram of Analyzer 2 1 3 4 5 6 RE V ECN 7 8 DESCRIPTION DRAW CHECHK APPR DATE D D Power supply board Heater voltage selection board C C Reagent refrigeration board Main Control Board Power drive board AD collection board B B APPROVALS MINDRAY DATE DESIGN CHECK A TITLE Host (220V) cabling diagram CHECK File: Bytes: Date: Time: CONFIDENTIAL DISCLOSURE: This set of drawing(s) and all it's intellectual property rights (including copyright) subsisting herein are property of Shenzhen Mindray Bio-medical Electronics Co.,Ltd. No use, copies or reproductions should be made of this drawing or any part(s) thereof for whatever purpose nor shall any information, data, calculations, or other contents contained in this drawing be disseminated without prior written permission of Shenzhen Mindray Bio-medical Electronics Co.,Ltd. Software & Rev: Microsoft office Visio 2003 1 BS-240 Chemistry Analyzer Service Manual 2 3 4 5 CHECK DWG NO. R&D SHEET 1 CHIEF ENG. 6 BA10 OF9 A REV 1.1 SIZE A4 7 6-35 6 Hardware Circuits 6.6 Wiring Diagram of Analyzer 6 Reserved valve Reserved pump Water inlet valve Heater under reaction carousel Heater temperature protection switch under reaction carousel Heater temperature protection switch on reaction carousel Heater on reaction carousel Wash solution temperature protection switch Deionized Water Heater 5 4 33PV3CTRL PV3CTRL 99yellow yellow 2 1 77black black 3 8 4 77GND GND 66yellow yellow 9 88220VCTRL3 220VCTRL3 10 33red red 88yellow yellow 10 10red red 11 12 33 34 8 7 6 5 22black black 12GND 12GND 66 88blue blue 10 10black black 5 2 6 3 5V 5V 11 20 20red red 6 19 19red red GND GND77 12V 12V 44 99black black 17 17yellow yellow 12GND 12GND 88 77blue blue 15 15orange orange 24V 24V 11 24V 24V 22 8 14 14orange orange GND GND 66 55white white GND GND 55 44white white B B12V B12V 33 12 12yellow yellow 55 B12V B12V 22 44 GND GND 66 GND GND 55 13 13yellow yellow 33blue blue 33 22 11 J1 J1 LL 11brown brown 7 9 3 10 2 5 1 1 2 N 33light lightblue blue N 3 AC ACpower powersupply supplycables cablesfrom from power powerswitch switchto toPFC PFCboard board and and24V 24Vboard board 11 1 12 2 13 3 14 4 Power Power supply supply board board C 1 1 12Vout 3 blue NO 2 3 C + 2 15 5 16 6 17 7 18 8 19 9 20 10 black Analyzer Switch GND B 22blue blue GND GND 44 66white white C12V C12V11 16 16orange orange ISE ISE power power supply supply PCBA PCBA ISE ISEmodule module 1124V 24V 22GND GND 3 1 33yellow yellow12V 12V 2 2 11blue blueGND GND 1 J2 J2 J1 J1 M32: M32:lamp lamp 12VGND 12VGND LAMP12V LAMP12V A 6 Refrigeration Refrigerationboard board12V 12Vpower powercord cord J86 J86 66 5 J2 5V 5V 33 J17 J17 7 4 Drive Driveboard boardpower powercord cord BA10-20-78108 BA10-20-78108 Power Power drive drive board board Refrigeration Refrigerationboard board12V 12V power powerpatch patchcord cord 3 J4 J4 12 12black black 18 18yellow yellow 4 2 11 11yellow yellow Main Maincontrol controlboard board 12V&5V 12V&5Vpower powercord cord 12V 12V 33 3 3 1 J14 J14 66220VCTRL2 220VCTRL2 GND GND 44 5 7 D LLbrown brown 44red red 1 1 2 11blue blue 99LAMPCTRL LAMPCTRL 12 12GND GND 2 1 6 55220VCTRL1 220VCTRL1 15 15GND GND C 5 44GND GND 20 20GND GND 4 J5 J5 55blue blue 22PV2CTRL PV2CTRL 12 Main Maincontrol controlboard boardcontrol controlsignal signal cable cable 11PV1CTRL PV1CTRL J2 J2 Housing Housingground ground Main Power Switch M07-00061S--- Power jack Main Main control control board board J16 8 7 Cables Cablesfrom frompower powerjack jack to topower powerswitch switch EEyellow yellowgreen green NNlight lightblue blue J3 J3 D 4 Wash Solution Heater 3 Dionized water temperature protection switch 2 1 11black black 22black black MINDRAY 11 11red red 11red red File: Bytes: Date: Time: Software & Rev: Microsoft office Visio 2003 1 6-36 2 3 4 5 6 A Power Supply System Cabling Diagram TITLE: 1.1 DWG NO. REV SHEET 2 OF 9 SIZE A4 7 BS-240 Chemistry Analyzer Service Manual 6 Hardware Circuits 6.6 Wiring Diagram of Analyzer 5 GND 4 8 34 5 7 33 11red red 11 11red red 11 yellow 88yellow 345V 5V 34 6 3 335V 5V 33 4 1 345V 5V 34 red 66red yellow 88yellow 220VCTRL3 88220VCTRL3 2 335V 5V 33 3 2 1 10 yellow 77yellow GND 77GND 9 8 blue 99blue black 11black 220VCTRL1 55220VCTRL1 7 5 6 red 44red blue 33blue black 22black yellow 55yellow GND 44GND Maincontrol controlboard boardcontrol controlsignal signal Main cable cable J16 1 2 3 PV3CTRL 33PV3CTRL 4 6 4 PV1CTRL 11PV1CTRL 4 GND 6 12GND 5 Main control board 12V&5V power cord 3 J11 PV2CTRL 22PV2CTRL 2 1 3 4 4 red 1 blue 3 12V 2 black 3 yellow 1 VCC 44 1 22 2 220VCTRL2 66220VCTRL2 2 red 4 44 3 3 22 1 48 47 1 50 49 46 45 44 43 42 41 38 37 40 39 36 35 34 33 28 27 32 31 30 29 26 25 24 23 18 17 20 19 9 4 16 15 7 8 14 13 12 11 2 3 48 47 50 49 42 41 46 45 38 37 44 43 40 39 36 35 28 27 34 33 32 31 30 29 26 25 18 17 22 21 24 23 20 19 16 15 7 9 4 8 14 13 12 11 J20 3 1 5 6 2 J5 J5 J4 2 J65 2 NC 3 NC 4 NC 5 NC 6 NC 7 NC 5 6 Cables from main control board to power drive board 1 GND D (Sample probe) level sense board PCBA 2 NC 3 NC 4 NC 5 NC 6 NC 7 NC 8 PUMP_FILL_IN 9 VAVLE_FILL_IN 10 PUMP_FILL_OUT 11 VALVE_WASH_INJECT1 12 VALVE_WASH_INJECT2 13 PREHEAT_REAG 14 STIR_DC 15 LAMP_CTRL 16 PWR_PV_WASH1 17 PWR_PV_WASH2 18 RELAY_T 19 GND 20 REAC-CLK 21 REAC-DIR 22 CLK_STIR_R 23 DIR_ STIR_R 24 DIR_WASH_UP 25 CLK_ WASH_UP 26 CLK_RESERVED_MOTOR1 27 DIR_RESERVED_MOTOR1 28 CLK_FILL_T 29 DIR_FIIL_T 30 CLK_RESERVED_MOTOR2 31 DIR_ RESERVED_MOTOR2 32 CLK_STIR_UP 33 DIR_STIR_UP 34 GND 35 CLK_SYR_MOTOR1 36 DIR_ SYR_MOTOR1 37 CLK_SYR_MOTOR2 38 DIR_ SYR_MOTOR2 39 CLK_RESERVED_MOTOR3 40 DIR_ RESERVED_MOTOR3 41 CLK_FILL_R 42 DIR_FILL_R 43 CLK_FILL_UP 44 DIR_ FILL _UP 45 CLK_FILL_SYR 46 DIR_FILL_SYR 47 GND 48 VCC 49 VCC 50 VCC 3 1 1 2 4 High-concentration waste sensor 2 yellow LEVEL1 3 J1 4 4 5 C 2 6 1 J66 1 1 white GND Diluted wash solution blank sensor Dionized water blank sensor 4 red LEVEL2 7 3 blue GND 8 NC NC 9 NC NC 10 NC NC J14 Power drive board BNC 22 21 BNC BNC 8 7 2 red 1 black Sample probe level detection cable BA30-20-06546 D 6 PROBE 4 C31: Sample probe 3 1 black 2 1 C J2 Main control board J39 6 8 2 4 6 8 10 3 5 7 1 3 5 7 9 7 CTS Yellow 6 8 TRIG Orange 9 9 NC 4 NC NC 5 NC 7 NC 8 NC 11 13 GND 10 DCP_CLK 25 GND 6 24 CH_A0 5 12 GND 4 NC GND 9 DCP_EN TXD232 23 CH_A1 11 DCP_DIN 3 22 CH_A2 CTS 8 9 NC 2 21 CH_A3 NC 6 RTS 7 7 Cable between main control board and AD board 1 NC 19 AD_CLK 6 NC 8 J23 RXD232 8 GND White 10 PC COM GND 5 6 AD_DIN Black 4 6 RTS NC 4 PC communications patch cord 20 AD_RC Green 3 Blue 2 5 GND 5 1 14 Red 2 RXD 3 TXD ISE module 2 15 NC 7 ISP_RESET 8 NC 15 14 P1 16 17 18 20 19 21 22 23 25 24 9 NC 1 2 3 4 5 6 7 8 9 10 10 11 12 B 13 NC NC 12 SHIELD NC 13 J21 J22 Red 2 4 NC Red 1 3 16 Black1 2 1 VCC 4 17 Reagent preheating sensor patch cord NC 1 RXD 2 TXD 3 5 18 19 3 red 1 NC 3 6 20 2 black Patch cord from main control board to ISE module 21 Black 1 red VCC 1 4 white PHO_FILL_T 2 2 black GND 3 3 green GND 4 1 red VCC 5 4 white PHO_FILL_TC 6 2 black GND 7 3 green GND 8 1 red VCC 9 PHO_WASH_UP 10 4 white 11 2 black GND 12 3 green GND 1 red VCC 13 4 white PHO_WASH_SUCKSYR 14 2 black GND 15 3 green GND 16 1 red VCC 17 18 4 white PHO_WASH_EJECT 2 black GND 20 3 green GND 1 red VCC 21 4 white PHO_RESERVED1 22 2 black 23 GND 3 green 24 GND VCC 25 4 white PHO_RESERVED2 26 27 2 black GND 28 3 green GND 1 red VCC 29 4 white PHO_RESERVED3 30 2 black 31 GND 3 green 32 GND 33 NC 34 NC 22 Reaction carousel temperature sensor cable + C E + C15:sample/reagent C carousel coder sensor E + C Standby E + C Standby E + Cuvette wash station C anti-collision E + Wash Station Home Position C E + C E + Wash Solution Syringe C Home Position E 23 2 PRTD_REF Cable 2 from main control board to position sensor C16:sample/reagent carousel home position sensor 7 J18 24 25 3 4 1 2 2 7 1 8 5 SENSOR 6 3 SHIELD 4 1 REF 2 33 AD collection board BA10-30-77759 NC 14 NC 15 3 34 31 2 32 29 1 30 25 27 Deionized water temperature sensor cable 26 28 23 3 24 21 SENSOR 22 19 3 20 17 2 18 15 REF 16 13 2 14 11 Wash solution temperature sensor cable 12 9 1 10 7 SHIELD 8 5 1 6 3 M29: Reagent Preheating Sensor 4 1 3 PRTD_T1 B 2 Dioni zed water syring e home positi on Cable 1 from main control board to position sensor 1 red VCC 1 + 4 white PHO_REAC_T 2 C 2 black GND 3 GND 4 E 3 green 1 red VCC 5 + 4 white PHO_REAC_TC 6 C Reaction Carousel Coder GND 7 - 2 black Sensor 3 green GND 8 E 9 1 red VCC + PHO_FILTER_MOTOR 10 4 white C C40: filter wheel 11 2 black GND home position sensor 12 GND E 3 green 1 red VCC 13 + C18: sample vertical 4 white PHO_FILL_UP 14 C movement motor GND 15 - 2 black home position sensor GND 16 E 3 green 1 red VCC 17 C17: sample probe rotation + 4 white PHO_FILL_R 18 C motor 19 GND - 2 black home position sensor 20 GND E 3 green 1 red VCC 21 + 22 PHO_FILL_SYR C19: sample syringe motor C 4 white 23 GND home position sensor - 2 black 3 green 24 GND E VCC 25 + 26 4 white PHO_STIR_UP C28: mixer vertical C 27 GND - 2 black movement motor 28 GND E 3 green home position sensor 1 red VCC 29 + 4 white 30 PHO_STIR_R C27: mixer rotation motor C 31 GND home position sensor - 2 black 3 green 32 GND E 33 NC 34 NC C13: reaction carousel home position sensor J13 18 AD_BUSY 33 5 GND 34 31 7 GND 32 29 4 15GND 30 25 27 16 VCC 26 28 23 3 VCC 24 21 17 15GND 22 19 1 +15V 20 17 14 +15V 18 15 2 -15V 16 13 15 -15V 14 11 Reagent bar code reader 12 9 C29: reaction carousel temperature sensor 10 7 8 3 8 5 9 2 6 3 10 1 4 1 11 2 2 12 13 3 J12 J7 J6 1 J5 MINDRAY A A TITLE: File: Bytes: Date: Time: Main control board cabling diagram BA10 DWG NO. SHEET 3 OF 9 Software & Rev: Microsoft office Visio 2003 1 BS-240 Chemistry Analyzer Service Manual 2 3 4 5 6 REV SIZE A2 7 6-37 1.1 6 Hardware Circuits 6.6 Wiring Diagram of Analyzer 5 3 1 8 6 4 2 P350 3 yellow 4 red 12V 4 VCC 3 1 blue 2 black GND 7 Power drive board power cord 2 white 12GND 8 3 orange 1 white P403 24V 2 GND 6 2 3 4 5 6 7 8 3 1 B 3 blue 1 red 2 blue GND 5V DGND 12V GND 24V J8 1 2 J13 3 4 J20 3 25 26 31 32 33 34 29 30 35 36 37 38 39 40 41 42 6-38 4 5 6 7 8 9 4 C 11 12 13 14 15 16 17 18 19 20 21 22 J40 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 B 41 42 43 44 45 46 47 48 49 50 2 blue 1 red 1 red 3 yellow 4 yellow MINDRAY TITLE: BA23 SHEET 4 OF 10 3 4 5 6 A Power drive board connection diagram DWG NO. 2 2 3 1 3 blue 2 3 4 4 white 3 6 white 1 red 1 red 1 6 2 blue 3 blue 3 4 Mixer vertical movement motor cable 4 white 3 yellow 1 6 white 2 4 yellow 3 4 6 1 red 4 Reaction carousel motor cable 1 red 1 2 blue 3 3 yellow 1 3 blue 2 Software & Rev: Microsoft office Visio 2003 1 1 J14 1 J16 4 yellow 3 4 6 white Mixer rotation motor cable 4 white 4 6 1 red 1 red 1 red 1 2 blue 2 blue 3 blue 2 1 red 4 white 3 J10 1 4 1 3 yellow 2 4 yellow 4 1 Reagent carousel motor cable 1 red 1 red 3 J4 6 white 2 blue 3 blue 3 4 6 4 white 3 yellow 1 6 white 2 4 yellow 6 4 Wash solution syringe motor patch cord 3 J15 3 yellow 4 4 Time: 1GND 2 NC 3 NC 4 NC 5 NC 6 NC 7 NC 8 9 10 11 12 13 14 15 37 38 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 16 17 39 40 41 42 43 44 45 46 47 48 49 49 50 J12 3 blue 1 red 3 1 red 4 3 1 45 46 1 GND 2 NC 3 NC 4 NC 5 NC 6 NC 7 NC 8 PUMP_FILL_IN 9 VAVLE_FILL_IN 10 PUMP_FILL_OUT 11 VALVE_WASH_INJECT1 12 VALVE_WASH_INJECT2 13 PREHEAT_REAG 14 STIR_DC 15 LAMP_CTRL 16 CLK_WASH_SYR 17 DIR_WASH_SYR 18 RELAY_T 19 GND 20 REAC-CLK 21 REAC-DIR 22 CLK_STIR_R 23 DIR_ STIR_R 24 CLK_RESERVED_MOTOR1 25 DIR_RESERVED_MOTOR1 26 CLK_WASH_UP 27 DIR_WASH_UP 28 CLK_FILL_T 29 DIR_FIIL_T 30 CLK_RESERVED_MOTOR2 31 DIR_ RESERVED_MOTOR2 32 CLK_STIR_UP 33 DIR_STIR_UP 34 GND 35 CLK_SYR_MOTOR1 36 DIR_ SYR_MOTOR1 37 CLK_RESERVED_MOTO4 38 DIR_RESERVED_MOTO4 39 CLK_RESERVED_MOTOR3 40 DIR_ RESERVED_MOTOR3 41 CLK_FILL_R 42 DIR_FILL_R 43 CLK_FILL_UP 44 DIR_ FILL _UP 45 CLK_FILL_SYR 46 DIR_FILL_SYR 47 GND 48 VCC 49 VCC 50 VCC 47 48 Cuvette wash station motor 6 white 4 white 2 blue 4 yellow 3 blue 6 2 3 43 44 1 1 3 yellow 3 4 white 4 yellow J11 6 white 4 Bytes: 23 24 27 28 6 Date: 4 21 22 Exterior Wash Syringe File: 9 19 20 Power drive board 1 red A 8 4 3 yellow 7 1 4 yellow 6 2 4 Reagent probe rotation motor cable 6 white 4 white 5 17 18 2 blue 1 red 4 11 12 J5 6 2 blue 1 red 2 3 3 1 1 15 16 3 yellow Cables from main control board to power board J6 4 3 4 D 1 GND 24V 2 J9 J17 P35 2 0 3 Main control board 7 2 1 2 GND 5 2 red 1 black 4 red 6 red 1 1 Bla ck R 2 e d 4 3 1 3 yellow 2 1 2 1 blue 2 red 3 4 3 4 orange M29: Reagent preheating block M19: Mixer DC motor 1 Patch cord from drive board to three-probe board preheater 2 red Patch cord from drive board to three-probe board mixer motor 1 1 blue 5 yellow 5 6 4 J7 4 7 8 Power supply 12&5V board Power supply 24V board 13 14 2 6 Filter wheel motor patch cord 00 4 white 6 white P01: interior pump 1 red 2 2 blue 1 red J24 8 7 1 3 blue 1 SV02: interior valve 3 6 1 3 yellow 2 4 4 yellow SV03: Sample Probe Exterior Valve 6 1 red Reagent probe vertical movement motor cable 6 white 4 white 2 C 1 red 2 1 2 J22 1 1 P04: phase 4 aspiration pump 2 blue P03: phase 2/3 aspiration pump 3 blue 2 3 P02: phase 1 aspiration pump 3 yellow 2 4 yellow SV05: Dionized water dispense valve 4 SV06: Wash Solution Dispense Valve 4 white 5 1 + 1 6 white SV04: Mixer Exterior Valve Syringe motor cable 6 2 2 - 1 D 4 24V 1 3 5 blue 2 1 7 BS-240 Chemistry Analyzer Service Manual REV 1 SIZE A4 6 Hardware Circuits 6.6 Wiring Diagram of Analyzer 2 1 3 4 5 6 8 7 D AGND_SHILED -15V AGND 5 6 AGND 4 3 +15V 2 Cables from AD collection board to preamplifier board 1 340nm Preamplifier board BA10-30-77760 D C C 2 4 6 1 3 5 J1 AD collection board 15 14 16 17 18 20 19 21 22 23 25 24 P1 B 2 4 3 1 5 6 7 8 9 10 11 12 13 B 1 +15V 14 +15V 2 -15V 15 -15V 3 VCC 16 VCC 4 15GND 17 15GND 5 GND 18 AD_BUSY 6 AD_DIN 19 AD_CLK 7 GND 20 AD_RC 8 GND 21 CH_A3 9 DCP_EN 22 CH_A2 10 DCP_CLK 23 CH_A1 11 DCP_DIN 24 CH_A0 12 GND 25 GND 13 GND Cable between main control board and AD board 15 14 J39 1 2 16 3 17 4 18 5 20 19 6 7 21 8 22 9 23 10 25 24 11 12 13 MINDRAY Main Control Board A TITLE: File: Bytes: Date: Time: DWG NO. 1 BS-240 Chemistry Analyzer Service Manual 2 BA10 SHEET 5 OF 9 Software & Rev: Microsoft office Visio 2003 3 4 5 6 A AD collection board cabling diagram 7 6-39 REV 1.1 SIZE A4 6 Hardware Circuits 6.6 Wiring Diagram of Analyzer 2 1 3 4 5 6 8 7 M42-M43: refrigeration 92 fan C35: reagent refrigeration temperature sensor J81 1 J79 1 2 J80 1 2 J78 1 2 J82 1 2 J83 1 2 GND Demisting temperatur e protection switch Demisting heater D J84 1 2 J15 3 1 2 4 +5V 2 2 1 J85 1 GND Demisting heater cable reserved 12V GND 12V 2 12V Reagent refrigeration/radiating fan cable 2 1 GND GND 12V 12V Rear panel fan patch cord 2 2 12V 1 1 GND J88 2 12V Reagent refrigeration temperature sensor patch cord 1 GND GND D 3 4 Power supply 12&5V board J89 J86 PELTIER cable 1 C 1 12V 12V 1 1 4 5 12V GND 2 5 4 12V GND 3 6 M30-M31:PELTIER 12VFANGNDFAN 2 GND 2 COOLER Reagent refrigeration board PELTIER cable 2 1 12V 1 blue 2 blue 3 whit e 4 yello w 5 6 3 12V 1 2 yello w 6 2 GND COOLER 2 1 oran ge Refrigeration board 12V power patch cord Cable between reagent refrigeration board and PCB fan GND 12V 12VFAN FAN3 FAN GND GNDFAN FAN4 12VFAN 12V FAN GNDFAN Lamp box radiating fan patch cord GND B 1 2 J77 1 2 1 J87 2 6 2 3 Refrigeration board 12V power cord 3 5V J76 FAN 2 GND 1 5 1 Feedback patch cord from main control board to refrigeration board fan J75 C 4 B 12V 3 GND 2 2 12V 1 12V J71 1 GND 3 GND 2 12V 1 GND J72 1 GND 3 2 12V 2 12V 1 Temperature control chamber fan patch cord J73 GND 3 12V 2 1 GND 1 2 12V J74 P353 J10 Power drive board radiating fan Lamp box fan cable Temperature control chamber fan cable Main control board BA30-30-77755 MINDRAY A File: Bytes: Date: Time: DWG NO. 1 6-40 A-BA20-30-75232 SHEET 6 OF Software & Rev: Microsoft office Visio 2003 2 3 4 5 6 A Reagent refrigeration board cabling diagram TITLE: 7 REV SIZE A4 7 BS-240 Chemistry Analyzer Service Manual 1.1 7 Hydropneumatic System BS-240 Chemistry Analyzer Service Manual 7-1 7 Hydropneumatic System 7.1 Overview 7.1 Overview The major function of the hydraulic part is: To provide deionized water for washing interior/exterior of probes and mixers via the water supply unit. To provide deionized water and wash solution for washing cuvettes via the cuvette wash station. To provide the whole unit with deionized water via the deionized water tank and to discharge waste liquid via the drainage module. This chapter describes the working principles and repairing methods of the BS-240's hydropneumatic system. 7-2 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.2 Fluidic Diagram 7.2 Fluidic Diagram Figure 7.1 Fluidic Diagram BS-240 Chemistry Analyzer Service Manual 7-3 7 Hydropneumatic System 7.2 Fluidic Diagram Tube label SV02 T104 T103 SV03 T221 D32 D07 SY03 T105 D05 Deionized water T10 6 D06 T10 7 MIX NS Low-concentration waste Reagent Reaction Carouse Carouse High-concentration waste Optional ZQ01 ZQ11 T61 3 D33 D34 SI1 T211 ZQ31 T222 SI2 SI3 SV04 D31 ZQ21 D22 D23 T212 SI4 W01 W11 W21 T312 W31 T30 ZZ23 5 SV07 W02 T52 1 T511 ZZ22 T30 4 ZZ21 T311 W22 ZZ31 T303 T502 T204 ZQ02 P04 P03 P02 ZQ33 ZQ14 ZQ03 ZZ24 ZZ08 ZZ32 T70 2 ZZ34 ZZ33 SV05 P01 T612 ZQ32 ZQ13 D24 D21 ZQ12 T61 1 T404 T405 1 ZZ05 2 3 ZZ07 D04 T602 SV06 ZZ12 D16 CAN02 ZZ04 ZZ06 ZQ15 ZZ11 T203 D11 T10 2 T202 D15 T201 T301 T302 T40 2 T40 3 SY02 D12 SV01 D14 D13 D03 CAN01 W03 SY01 ZZ03 Panel D02 ZZ02 T401 T101 FL01 FL02 D01 T60 1 ZQ16 V01 T701 ZQ04 FS03 ZZ01 FS01 7-4 T501 W04 FS02 V02 V03 (optional) BS-240 Chemistry Analyzer Service Manual V04 7 Hydropneumatic System 7.2 Fluidic Diagram Fluidic Port Panel Dil Detergent inlet DI water inlet Dionized water low level sensor Low-conc. waste outlet1 Low-conc. waste outlet2 High-conc. waste outlet Highconcentration waste high level sensor Diluted wash solution low level sensor FL02 FL04 FL03 FL01 Dil Detergent tank (CAN2) BS-240 Chemistry Analyzer Service Manual High-conc. waste tank (CAN10) 7-5 7 Hydropneumatic System 7.2 Fluidic Diagram Analyzing Unit Sensor Tube Lowconcentration waste Highconcentration waste Wash solution tank DI water tank Table 7.1 Silicone tube lengths and identifiers The tube length tolerance of the tube is not specified: when the tube length ≤ 100mm, the tolerance is ±5mm; when the tube length > 100mm, the tolerance is ±10mm. 7-6 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.2 Fluidic Diagram Tube label Rubber hose Type Part No. Actual length Number of Labels Tube Label type Remark D03 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 220 1 PVC 8mm hose Assembling fluidic outlet assembly D04 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 200 1 PVC 8mm hose Assembling fluidic outlet assembly D05 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 350 1 PVC 8mm hose D06 Rubber hose .PTFEID1.5mmXOD2.5mm×100M M6G-020049--- 320 / / D07 Rubber hose .PTFE,0.040"IDX0.066"OD 0040-10-32301 2300 / / D11 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 400 1 PVC 8mm hose Assembling fluidic outlet assembly D12 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Assembling fluidic outlet assembly D13 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 500 1 PVC 7mm hose D14 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Exterior wash tubing D15 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 40 1 PVC 8mm hose Exterior wash tubing D16 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 30 1 PVC 8mm hose Exterior wash tubing D21 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 60 1 PVC 8mm hose Exterior wash tubing D22 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Exterior wash tubing D23 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Assembling instrument body assembly BS-240 Chemistry Analyzer Service Manual Assembling reagent preheating assembly 7-7 7 Hydropneumatic System 7.2 Fluidic Diagram D24 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 900 2 PVC 8mm hose Assembling instrument body assembly D31 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 60 1 PVC 8mm hose Exterior wash tubing D32 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Exterior wash tubing D33 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Assembling instrument body assembly D34 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 800 2 PVC 8mm hose Assembling instrument body assembly ZZ03 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 560 2 PVC 8mm hose Assembling fluidic outlet assembly ZZ04 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 360 1 PVC 8mm hose ZZ05 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose ZZ06 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 200 1 PVC 7mm hose ZZ07 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 90 1 PVC 7mm hose Assembling wash tube ZZ08 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 220 1 PVC hose 1.5mm Assembling wash tube ZZ11 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 160 1 PVC 8mm hose Exterior wash tubing ZZ12 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 200 1 PVC 8mm hose Dionized waster preheater tubing ZZ21 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 50 1 PVC 8mm hose Dionized waster preheater tubing ZZ22 Rubber 65,Tygon M90-100071--- 50 1 PVC 7mm hose Dionized waster preheater tubing 7-8 hose .3/32"X5/32",ND-100- BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.2 Fluidic Diagram ZZ23 Rubber 65,Tygon ZZ24 hose M90-100071--- 120 1 PVC 7mm hose Assembling wash tube Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 260 1 PVC hose 1.5mm Assembling wash tube ZZ31 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 200 1 PVC 8mm hose Dionized waster preheater tubing ZZ32 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 50 1 PVC 7mm hose Dionized waster preheater tubing ZZ33 Rubber 65,Tygon hose .3/32"X5/32",ND-100- M90-100071--- 120 1 PVC 7mm hose Assembling wash tube ZZ34 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 260 1 PVC hose 1.5mm Assembling wash tube ZQ01 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 350 1 PVC hose 1.5mm Assembling wash tube ZQ02 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 400 1 PVC 8mm hose Assembling wash tube ZQ03 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 400 1 PVC 8mm hose Assembling fluidic outlet assembly ZQ11 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 280 1 PVC hose 1.5mm Assembling wash tube ZQ12 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 100 1 PVC hose 1.5mm Assembling wash tube ZQ13 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 400 1 PVC 8mm hose Assembling wash tube ZQ14 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 100 1 PVC 8mm hose Assembling fluidic outlet assembly ZQ15 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 300 1 PVC 8mm hose Assembling fluidic outlet assembly ZQ21 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 280 1 PVC hose 1.5mm BS-240 Chemistry Analyzer Service Manual .3/32"X5/32",ND-100- Assembling wash tube 7-9 7 Hydropneumatic System 7.2 Fluidic Diagram ZQ31 Rubber hose .1.6*3.2mm TPU polyether 082-002375-00 420 1 PVC hose 1.5mm Assembling wash tube ZQ32 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 400 1 PVC 8mm hose Assembling wash tube ZQ33 Rubber hose .3.2*6.4mm TPU polyether 082-002374-00 140 1 PVC 8mm hose Assembling fluidic outlet assembly W01 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 240 / / Assembling instrument body assembly W02 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 180 / / Assembling instrument body assembly W03 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 150 / / Assembling fluidic outlet assembly W11 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 200 / / Assembling instrument body assembly W21 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 110 / / Assembling instrument body assembly W22 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 140 / / Assembling instrument body assembly W31 Rubber hose .Φ9.525XΦ15.875 natural color PVC 55~60° 082-000384-00 150 / / Assembling instrument body assembly 7-10 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.3 Principles of Hydropneumatic System 7.3 Principles of Hydropneumatic System The hydropneumatic system of the BS-240 is composed of the probe wash module, cuvette wash station, and water supply/drainage module. See the figures below: Figure 7.2 Principles of Hydropneumatic System Hydropneumatic System Probe and Mixer Wash Module Cuvette Wash Module Inlet/outlet module Waste drainage Inlet Reaction cuvette aspiration Reaction cuvette dispensing Mixer cleaning Probe Wash 7.3.2 Probe wash module The probe/mixer wash module consists of: 1 sample probe, 1 mixer, 2 syringes, and 1 diaphragm pump. The sample syringe SY03 is 500μl. The probe and the mixer exterior are washed by a 10ml syringe SY01. The probe interior is washed by the diaphragm pump P01. This module is used for fixed-quantity sampling and two-probe cleaning. Fixed-quantity sampling is realized through cooperation of the syringes, the sampling solenoid valve SV02 and the sample probe NS with the aim of delivering fixed-quantity reagent and sample. The sample probe interior is washed by a diaphragm pump. The probe and the mixer exterior are both washed by a 10ml syringe. The start and end of the wash action are controlled by the exterior wash solenoid valve. A deionized water T-valve controls the dispensing of the syringe. The volume limiter of the syringe controls the volume of the syringe. The schematic diagram of the probe wash module is shown below. BS-240 Chemistry Analyzer Service Manual 7-11 7 Hydropneumatic System 7.3 Principles of Hydropneumatic System Figure 7.3 Schematic diagram of probe wash module SV02 T10 6 D06 T104 T103 SY03 T105 D05 D07 SV03 T10 7 MIX NS T222 T221 D32 D33 D34 SV04 D22 T211 D31 D23 T212 D24 W01 D21 P01 T204 D04 D16 ZZ11 T203 D11 T10 2 T202 D15 T201 D12 SV01 D14 D13 D03 SY0 1 Panel D02 T101 FL01 7.3.3 Cuvette Wash Module The cuvette wash unit is divided into dispense module and aspirate module, which cooperate with each other to clean the reaction cuvettes for 4 phases, making repeated use of cuvettes possible. The dispense module uses a cuvette wash syringe SY02 (2.5ml) to provide diluted wash solution (alkaline) in phase 1. In phases 2 to 3, the cuvette wash syringe SY01 (10ml) provides deionized water. In phase 4, no liquid is injected. 7-12 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.3 Principles of Hydropneumatic System The diluted wash solution provided for phase 1 is obtained by diluting the concentrated wash solution with deionized water at the ratio of 1:10. Samples are diluted manually. The dispense action is controlled by the exterior wash solenoid valve and adjusted by the syringe volume limiter. The dispensed volume is the same in phase 1 to phase 3. A diaphragm pump is used to aspirate waste, so as to discharge waste fluid and wipe the cuvettes after cleaning. High-concentration waste is generated in phase 1, and lowconcentration waste during phases 2-4. Wipe blocks are provided to absorb the remaining wash solution inside cuvettes during phase 4. When aspirating fluid, the phase-4 wash probe assembly lowers into reaction cuvettes, and then waste pump is opened so that the waste is aspirated into the external waste container. The schematic diagram of the module is shown below. Figure 7.4 Schematic diagram of cuvette wash unit ZQ01 ZQ11 ZQ21 ZQ31 T613 SI1 SI2 SI3 ZQ12 SI4 T611 T612 ZQ32 ZQ13 P04 T312 P02 ZZ34 ZZ33 SV05 P03 T702 ZQ02 ZZ23 T305 SV07 ZZ24 ZZ08 ZZ32 ZQ33 ZQ14 ZZ22 T311 ZQ03 T304 ZZ21 ZZ31 T303 T404 T405 1 ZZ05 ZZ07 2 3 T602 SV06 ZZ12 CAN02 ZZ04 ZZ06 ZQ15 T203 T302 T301 T202 D15 T201 D12 D14 SY02 SV01 T403 T402 CAN01 D13 W03 SY01 ZZ03 7.3.4 Inlet/outlet module The water supply/drainage module includes the water supply part and waste drainage part. BS-240 Chemistry Analyzer Service Manual 7-13 7 Hydropneumatic System 7.3 Principles of Hydropneumatic System Inlet deionized water is provided from the outside of the instrument, and is conveyed by the interior wash pump and the deionized water syringe to other modules. Inlet diluted wash solution is provided by an external diluted wash solution tank outside the instrument, and is conveyed by the wash solution syringe to phase 1 of wash, as detailed below: detecting deficient fluid level of dionized water, and alarming. detecting deficient fluid level of diluted wash solution, and alarming. filtering deionized water and diluted wash solution, and performing filtering before the inlet deionized water and the inlet diluted wash solution outside the whole unit are distributed to each unit. The drainage module allows discharging high-/low-concentration waste separately. The high-concentration waste is discharged by the waste pump to the external highconcentration waste tank; the low-concentration waste is discharged by the waste pump of the low-concentration waste tank outside the instrument (and may be discharged via sewers directly). The high-concentration waste tank is capable of detecting the fluid level. The high-concentration waste is produced during cuvette cleaning in phase 1. Low-concentration waste comes from the following sources: Cleaning interior/exterior of sample probe and reagent probes Cleaning exterior of sample mixer and reagent mixer Auto cuvette cleaning in phase 2-4 Condensate water of the reagent carousels Overflowed fluid of the reaction carousel The first three are the major sources and account for over 99% of low-concentration waste. The fluidic diagram of the module is shown below. 7-14 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.3 Principles of Hydropneumatic System Figure 7.5 Fluidic diagram of water supply/drainage module Reagent Carouse MIX NS Reaction Carouse T602 W01 W11 W21 W31 ZQ15 T521 W02 T511 W22 T502 W03 ZZ03 D02 ZZ02 T101 FL01 T401 FL02 D01 T501 W04 T601 ZQ16 FS03 ZZ01 FS01 V01 T701 ZQ04 FS02 V02 V03(选配) V04 7.3.5 Other The Hydropneumatic system of BS-240 contains 8 external interfaces, as shown in the figure below, where, 3 fluidic outlets: used for discharging high-concentration waste, low-concentration waste 1, and low-concentration waste 2. 2 fluidic inlets: used for supplying deionized water and diluted wash solution. 3 control cable connectors: used for connecting the high-concentration waste level sensor, deionized water level sensor, and diluted wash solution level sensor. BS-240 Chemistry Analyzer Service Manual 7-15 7 Hydropneumatic System 7.3 Principles of Hydropneumatic System Figure 7.6 External fluidic interfaces of the BS-240 Interfaces for fluid connection No. Name 1 dionized water 2 diluted wash solution 3 high concentration waste 4 low concentration waste 1 5 dionized water sensor 6 diluted wash solution sensor 7 high concentration waste sensor 8 low-concentration waste 2 Water Waste Wash solution Sensor 8 4 7 3 6 2 5 1 Legend 7-16 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.4 Introduction of Fluidic Actions 7.4 Introduction of Fluidic Actions 7.4.1 Fluidic Initialization The specific procedure is described below: 1 Power off all the pump and valve and they return to the home position. 2 The waste pump discharges waste. 3 The floater status of the deionized water tank, diluted wash solution tank, and high concentration waste tank are checked. If related alarm is removed, the system moves to the next step. If not, the system gives out the alarm and the fluidic initialization fails. 4 Cleaning Probes/Mixers Exterior is performed. 5 The probe and mixer exterior wash syringe is reset; and phase1 wash syringe is reset. NOTE: The fluidic initialization is one part of the startup , system recovery and system wakeup procedure. 7.4.2 Priming Wash Station The specific procedure is described below: 1 Open wash waste pump P02~P04. 2 Let the wash station move down to the cuvette bottom. 3 Prime the wash station cyclically 3 times. 4 After completion of priming, close the wash solenoid valve, reset the wash syringe, let the wash station move to the home position, and close the waste pump P02~P04. BS-240 Chemistry Analyzer Service Manual 7-17 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components 7.5 Removing and Reinstalling Hydropneumatic Components This Section describes removing and installation methods of the Hydropneumatic components, and provides schematic diagrams and pictures of components for service engineers to refer. 7.5.1 Overview When an alarm is displayed on the operating software of BS-240, service engineers can analyze the instrument status and finally locate the failure source, and if necessary, replace the failed part or component. Generally, service engineers are not recommended to remove the electric devices, such as pumps, solenoid valves, clog detection device, syringes, etc. Only when both hardware and software are confirmed normal but the Hydropneumatic alarm still remains are service engineers suggested to remove the relevant device and then analyze or replace it. Prior to removing Hydropneumatic components, make sure that all pumps and solenoid valves of the Hydropneumatic system have been turned off, and both the analyzer power switch and main power switch have been placed to the OFF position. Table 7.2 Necessary tools for removing/installing Hydropneumatic components Name 7-18 Quantity Cross head screwdriver 1 Hexagon screwdriver 1 Flathead screwdriver 1 Strap Several Diagonal pliers 1 Tube cutter 1 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components 7.5.2 Probe and Mixer Wash Module Figure 7.7 Assembly drawing of whole unit –Probe and mixer wash module-front view Sample probe 500μl syringe Mixer Probe wash well Mixer wash well BS-240 Chemistry Analyzer Service Manual 7-19 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Figure 7.8 Assembly drawing of whole unit -- Probe wash module-rear view Deionized water inlet T-valve Probes exterior wash valve Mixer Exterior Wash valve Deionized Water Syringe Probe interior wash valve Probes Interior wash pump Table 7.3 List of materials No. FRU No. or Part No. Part Name Remark 1 801-BA23-00025-00 Syringe 500ul(ZDV) FRU 2 082-002435-00 Two-position two-way valve FRU/probe interior wash valve 3 BA30-21-15311 KNF diaphragm pump assembly FRU/probe interior wash pump 4 115-011901-00 10 ml screw rod driving syringe assembly (FRU) FRU/deionized water Syringe 5 082-002265-00 Vertically moving 3-way LVMK202-6J 90kPa~0.2MPa 12VDC FRU/ valve 6 082-002264-00 Vertically moving 2-way LVMK21-6J 90kPa~0.2MPa 12VDC FRU/probe exterior wash valve 7 M6G-020049--- Rubber hose .PTFEID1.5mmXOD2.5mm×100M FRU 8 0040-10-32301 Rubber hose .PTFE,0.040"IDX0.066"OD FRU 9 082-002374-00 Rubber hose polyether FRU 10 801-BA80-00009-00 Wash well shell FRU 11 801-BA40-00221-00 Wash well (mould MR61463) FRU 7-20 .3.2*6.4mm TPU deionized water inlet BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Removing/Reinstalling Wash Syringe Assembly When to do Analyze or replace the wash syringe when it has abnormal sound, worse precision, leaking, or other failures. Disassembling Steps 1 Remove the 4 M4×8 cross pan head screws used to fix the syringe shield and remove the shield. 2 Disconnect the syringe motor cable and sensor cable. 3 Remove the tube from the barbed connector of the syringe. 4 Remove the four M3×12 hexagon socket head cap screws with spring washer and remove the syringe. Installation Procedure 1 Install the 4 M3×12 hexagon socket head cap screws with spring washer and do not miss the shock pad . 2 Cut off a little from the tube's end and insert the tube to the barbed connector of the syringe. If a strap has been applied, restore it. 3 Connect the syringe motor cable and sensor cable. 4 Use 4 M4×8 cross pan head screws to fix the syringe shield on the right side plate. 5 Power on the analyzer, log on the software with service user account, and then select Utility -> Maintenance -> Maintenance -> Engineer -> Clear. If the interior wash syringe is replaced, select the "Replace Interior Wash Syringe" item. If the cuvette wash syringe is replaced, select the "Replace Cuvette Wash Syringe" item. Click OK to zero the use count of the relevant syringe. Figure 7.9 Details of wash syringe assembly Right baffle weld (BA49) Wash Syringe Assembly 115-013114-00 Syringe dust shield M4×8 cross pan head screw M3×12 hexagon socket cap head screw, with spring/flat washer BS-240 Chemistry Analyzer Service Manual 7-21 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Removing/Installing Solenoid Valves When to do When a solenoid valve cannot be turned on or has leaks, it needs to be removed and then analyzed or replaced. Disassembling Steps 1 Disconnect the solenoid valve’s power cord connector. 2 Mark the installation direction of the valve and the connection mode of the inlet and outlet tubes to prevent them from being confused. 3 Remove the tubes and apply strap on their openings to prevent liquid spraying. 4 Loosen the screws on the solenoid valve and then remove the solenoid valve. Installation Procedure 1 Check the installation direction according to the marks, connect the inlet and outlet tubes, and then tighten the tube clamps or straps. 2 Tighten the retaining screws to fix the solenoid valve. 3 Make sure that the power cord is connected to the correct positive and negative ends. Verification steps Refer to Section 8.9 Hydropneumatic Unit. 7-22 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Figure 7.10 Details of wash solenoid valve assembly Disassembling/assembling interior wash pump When to do The pump does not work, that is, there is no flow or pressure. The pump flow and pressure is low. The pump has leaks. Noise is produced when the pump is working. Disassembling Steps 1 Disconnect the pump’s power cord connector. 2 Mark the connection mode of the inlet and outlet tubes and prevent them from being confused. 3 Remove the tubes and apply straps on their openings or use a tank to hold water, so as to prevent liquid from entering the instrument. 4 Loosen the screws fixing the pump and then remove the pump. Installation Procedure 1 Connect the inlet and outlet tubes according to the marks and then tighten the tube clamps. BS-240 Chemistry Analyzer Service Manual 7-23 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components 2 Tighten the retaining screws to fix the pump. 3 Make sure that the power cord is connected correctly. Verification steps Refer to Section 8.9 Hydropneumatic Unit. NOTE: Exercise caution to avoid confusing the inlet tube with the outlet tube. Make sure that the power cord of the DC pump is connected to the correct positive and negative ends. Figure 7.11 Pump assembly structure Probes Interior wash pump 7-24 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components 7.5.3 Cuvette Wash Module Figure 7.12 Assembly drawing of whole unit -- Cuvette wash module –rear view Wash Solution Syringe Wash solution shift valve Phase-2 wash solution dispense valve Phase-3 wash solution dispense valve Deionized Water Syringe Wash preheating tank P02-P04 waste pump BS-240 Chemistry Analyzer Service Manual 7-25 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Figure 7.13 Assembly drawing of whole unit -- Cuvette wash module –front view Wipe block Cuvette wash station Wash probe assembly Table 7.4 List of materials No. FRU name Remark 2000-21-06116 Rotational diaphragm pump assembly Phase 1-8 Wash Waste Pump 082-002273-00 Soneloid cable Phase 3-6 cuvette solenoid valve wash 3 BA38-21-88190 WTB-3R-N4E T-valve cable Phase 1-2 cuvette solenoid valve wash 4 115-011901-00 10 ml screw rod syringe assembly Cuvette assembly 5 115-020528-00 Preheating assembly FRU 6 043-000422-00 Filter FRU 7 115-027921-00 Wash probe assembly FRU 8 041-005591-00 Wipe block FRU 9 BA38-30-88154 Wash phase probe assembly FRU 10 BA40-30-61934 Wipe phase probe assembly FRU 1 2 7-26 Part No. valve LVMK21-6J driving wash syringe BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Removing/Reinstalling Wash Syringe Assembly When to do Analyze or replace the wash syringe when it has abnormal sound, worse precision, leaking, or other failures. Disassembling Steps 1 Disconnect the syringe motor cable and sensor cable. 2 Remove the tube from the barbed connector of the syringe. 3 Remove the four M3×12 hexagon socket head cap screws with spring washer and remove the syringe. Installation Procedure 1 Install the 4 M3×12 hexagon socket head cap screws with spring washer and do not miss the shock pad . 2 Cut off a little from the tube's end and insert the tube to the barbed connector of the syringe. If a strap has been applied, restore it. 3 Connect the syringe motor cable and sensor cable. 4 Power on the analyzer, log on the software with service user account, and then select Utility -> Maintenance -> Maintenance -> Engineer -> Clear. If the cuvette wash syringe is replaced, select the "Replace Cuvette Wash Syringe" item. Click OK to zero the use count of the relevant syringe. Verification steps Refer to Section 8.9 Hydropneumatic Unit. Removing/Reinstalling Waste Pump When to do The pump does not work, that is, there is no flow or pressure. The pump flow and pressure is low. The pump has leaks. Noise is produced when the pump is working. Disassembling Steps 1 Disconnect the pump’s power cord connector. 2 Mark the connection mode of the inlet and outlet tubes and prevent them from being confused. 3 Remove the tubes and apply straps on their openings or use a tank to hold water, so as to prevent liquid from entering the instrument. 4 Loosen the screws fixing the pump and then remove the pump. BS-240 Chemistry Analyzer Service Manual 7-27 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Installation Procedure 1 Connect the inlet and outlet tubes according to the marks and then tighten the tube clamps. 2 Tighten the retaining screws to fix the pump. 3 Make sure that the power cord is connected correctly. Verification steps Refer to Section 8.9 Hydropneumatic Unit. NOTE: Exercise caution to avoid confusing the inlet tube with the outlet tube. Make sure that the power cord of the DC pump is connected to the correct positive and negative ends. Phase 1-3 waste pump Removing/Reinstalling Cuvette Wash Preheating Assembly When to do The heater is damaged. The temperature sensor is damaged. The temperature protection switch is damaged. The canister has leaks. The connectors are leaking. Disassembling Steps 1 7-28 Switch off the power supply of the whole unit. BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components 2 Open the rear panel. 3 Mark the connection modes of the following parts: inlet/outlet tubes, heater, temperature protection switch and temperature sensor for wash solution, and inlet/outlet tubes, heater, temperature protection switch and temperature sensor for deionized water. 4 Disconnect the connectors of the heaters, sensors and protection switches. 5 Loosen the two M4×8 hexagon socket head screws fixing the cuvette wash preheating assembly. 6 Apply straps on the tube openings to prevent liquid from flowing out. 7 Disconnect the inlet and outlet tubes. Installation Procedure 1 Prepare a new cuvette wash preheating assembly. 2 Connect the connectors according to the marks. 3 Connect the tubes according to the marks. 4 Use 2 M4×8 hexagon socket head screws to fix the preheating assembly. 5 Check if the installation is complete. 6 Restore the instrument. 7 Connect the power supply and perform fluidic prime before testing the system. Verification steps Refer to Section 8.9 Hydropneumatic Unit. NOTE: Operate carefully to prevent liquid from entering the instrument. Connect the tubes and connectors correctly. Do not confuse the heater and sensor channel of wash solution with those of cleaning fluid. BS-240 Chemistry Analyzer Service Manual 7-29 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components Figure 7.14 Structure of cuvette wash preheating assembly Wash solution outlet Wash solution heater/sensor Wash solution inlet Deionized water heater/sensor Deionized water inlet Deionized water outlet Removing/Reinstalling Filter When to do The filter should be maintained regularly. Disassembling Steps 1 The filter is located outside and under the inlet assembly. 2 Remove the filter tube and the filter, and then apply straps on the tube opening to prevent liquid inside the tube from spraying out. Installation Procedure 1 Install the filter tube and apply a strap. Verification steps Refer to Section 8.9 Hydropneumatic Unit. Figure 7.15 Assembly drawing of whole unit -- Acetabular filter Removing/Reinstalling Wash Probe Assembly When to do 7-30 BS-240 Chemistry Analyzer Service Manual 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components When fluid overflows the cuvettes or the dispensed cleaning fluid is less than the normal volume, probably a wash probe is clogged and needs to be removed and unclogged. Disassembling Steps 1 Manually loosen the retaining screws on the wash probe bracket. 2 Remove the wash probe assembly and place it in a container. 3 Turn on the dispense valves and vacuum valves, and then locate the clogged wash probe. 4 Use a needle to unclog the probe. Installation Procedure 1 Align the locating holes on the wash probe assembly with the stop studs on the bracket, and then slightly tighten the retaining screws. 2 Lower the wash probe assembly, use an alignment fixture or eyes to check the phase 4 wipe blocks corresponding to the cuvette wall, and then slightly adjust the wash probe assembly to prevent the wash probes from colliding with the cuvettes. 3 After finishing the second step, completely tighten the retaining screws on the wash probe assembly. Verification steps Refer to Section 8.9 Hydropneumatic Unit. NOTE: Operate carefully to prevent liquid from dropping into the reaction carousel and cuvettes. 7.5.4 External Modules External high-concentration waste tank cap assembly The external high-concentration waste tank cap assembly contains one FRU. Table 7.5 High-concentration waste tank cap assembly FRU No 1 FRU No. or Part No. 115-037348-00 Part Name Waste tank cap assembly Remark FRU External diluted wash solution tank cap assembly The external diluted wash solution tank cap assembly contains one FRU. Table 7.6 High-concentration waste tank assembly FRU No FRU No. or Part No. BS-240 Chemistry Analyzer Service Manual Part Name Remark 7-31 7 Hydropneumatic System 7.5 Removing and Reinstalling Hydropneumatic Components 1 115-037347-00 Wash solution tank cap assembly FRU External deionized water tank cap assembly The external deionized water tank cap assembly contains one FRU. Table 7.7 High-concentration waste tank assembly FRU No 1 7-32 FRU No. or Part No. 115-037347-00 Part Name Wash solution tank cap assembly Remark FRU BS-240 Chemistry Analyzer Service Manual 8 Alignment BS-240 Chemistry Analyzer Service Manual 8-1 8 Alignment 8.1 Basic Operation 8.1 Basic Operation 8.1.1 Utility - Maintenance Table 8.1 Utility - Maintenance No Description 1, 2 Maintenance and Diagnostics: used to maintain and diagnose the system at user end. 3 Alignment: used to align all mechanical positions, and the hydropneumatic, pyrology and ISE units. 4 Parameters: used to inquire and configure parameters, and to export parameters of each unit. 5 SPT: provides basic performance tests for the system, including photometer test, precision test (dye), precision test (weight), cuvette residue test, and carryover test. 6 Export Parameters: allows alignment parameters exporting mechanical position Figure 8.1 "Maintenance" screen 8-2 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.1 Basic Operation NOTE: Before replacing the main control board, you must export the alignment parameters to a local folder on the Parameter Configuration window. After finishing the replacement, import the parameters to the software. Otherwise, the system, when started up, will give an alarm indicating key parameters not configured and cannot be restored. Also change the parameter of fluidic prime status. 8.1.2 Alignment fixtures Table 8.2 Alignment fixtures No Code Name Image/Precision Quantity 1 BA24-J04 BA24 cuvette alignment fixture 2 2 BA10-J16 Reagent bottle center alignment fixture 3 3 BA48-J10 Sample position alignment fixture 3 4 BA60-J07 Pseudo probe 1 5 BA80-J20001 Alignment fixture block 1 6 BA48-J16 Mixer alignment lever 1 7 BA24-J05 BA24 mixer horizontal position alignment cuvette 1 BS-240 Chemistry Analyzer Service Manual 8-3 8 Alignment 8.1 Basic Operation No Code Name Image/Precision Quantity 8 BA40-J76 Mixer wash alignment fixture well 1 9 BA24-J01 Wash probe position alignment fixture 1 10 BA48-J12 Sample probe wash well alignment fixture 1 11 BA24-J03 Reagent carousel bar code reader alignment fixture 1 12 / Clearance gauge 0.5mm 1 8.1.3 Mechanical Reset Alignment methods and procedure: 8-4 1 Install the pseudo probe (BA60-J07), and block the collision sensor with folded thick paper. 2 Access the Alignment window and select Mech Reset to reset the mechanical parts of the system. BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.2 Alignment Procedure 8.2 Alignment Procedure Figure 8.2 Alignment Procedure Start Preparation Powering On Photoelectric unit adjustment Signal Collecting Position Adjustment Bar Code Unit (Optional) Bar code label printing and affixing Sample/reagent carousel unit parameter adjustment Reaction Carousel Circumferential Position Adjustment Setting Parameters of Probe Unit Aligning Horizontal Position from Probe to Reaction Carousel Sample Probe Rotary to SIC Height (Optional) Setting Parameters of Cuvette Wash Unit Adjust Horizontal Position from Wash Probe to Reaction Cuvette Adjusting Horizontal Position from Mixer to Reaction Carousel Preparations before Alignment Setting Pyrology Unit Parameter Setting Reaction Carousel Sensor Parameter Download program of upper layer/lower layer unit Photoelectric gain adjustment Adjusting Bar Code Reader Position Testing reagent bar code scanning stability Sample/Reagent Carousel Outer Ring Radial Position Sample/Reagent Carousel Middle Ring Radial Position Adjust Horizontal Position from Probe to Sample/Reagent Carousel Outer Ring Sample Probe Rotary to ISE (Optional) Setting Mixer Unit Parameter Fluidics Alignment Installation of Operating Software Sample Carousel Inner Ring Radial Position Adjust Horizontal Position from Probe to Sample/Reagent Carousel Middle Ring Aligning Horizontal and Vertical Positions from Probe to Wash Well Vertical Extreme Position from Probe to Reagent Bottle Adjust Horizontal Position from Probe to Sample/Reagent Carousel Inner Ring Sample Probe Cuvette Height Cuvette Wash Station Height Adjusting Mixing Position from Mixer to Reaction Carousel Checking Floater/ Valve Status Reaction carousel heating detection Adjusting Horizontal Position from Mixer to Wash Well Syringe Check Wash Solution Temperature Curve Fluidic Prime Reagent refrigeration adjustment Completing parameter configuration table and test record table End BS-240 Chemistry Analyzer Service Manual 8-5 8 Alignment 8.3 Photoelectric Unit 8.3 Photoelectric Unit Select Utility—>Maintenance—>Alignment—>Photoelectric unit. The window is shown below. 8.3.1 Photoelectric collection position adjustment Alignment index: The collected signal of 3# cuvette position is between two border lines. Alignment methods and procedure: 1 The reaction carousel location sensor is located on the right side of the whole unit. To perform alignment, the right cover needs to be opened. 2 Use an M3 hexagon wrench to loosen 3 screws that fasten the sensor. Put the sensor in the middle of the adjustable range, slightly tighten a screw, and then select Signal Collecting Position Adjustment to display the alignment window. 8-6 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.3 Photoelectric Unit Reaction carousel position sensor 3 Make sure the lamp has been turned on for over 1 minute. If not, wait for 1 minute, and then select "Start" to proceed to the next step. 4 A prompt box appears. Select "Start", and use a pipettor to add 200μl deionized water into the reaction cuvette indicated on the screen. Select "Next". 5 Observe the waveforms in the upper part of the window and check if the photoelectric signal meets the alignment requirement. See the figure below. If not, adjust the photocoupler. BS-240 Chemistry Analyzer Service Manual 8-7 8 Alignment 8.3 Photoelectric Unit To adjust the photocoupler, loosen the 3 M3 hexagon screws that fasten the photocoupler bracket. Use a wrench to hit the sensor gently to adjust its location, and the green box graph moves rightward. Adjust the sensor location counterclockwise, and the green box graph moves leftward. After the adjustment, tighten one screw slightly. Click "Realign", and the software performs tests and judgment again. If the photocoupler is in a proper position, select Done without any adjustment. 6 Tighten the sensor screws. 8.3.2 Photoelectric gain adjustment Index Setup 8-8 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.3 Photoelectric Unit Figure 8.3 Access the "Index Setup" screen: The water blank AD value of each channel is set to 48000-54000. The optical gain coefficient of all channels is generally set to 40 for a new instrument, and may be set to 30 for a client-side old instrument. The dark current of all channels is 1~200. Alignment methods and steps: 1 Select Photoelectric Gain Adjustment. 2 Make sure the lamp has been turned on for over 3 minutes, and then select "Done". If you do not click "Done", you need to wait for 3 minutes. 3 Add 200uL deionized water into cuvettes 1~5 in segment 1 according to step 4 on the screen. 4 The photoelectric gain is adjusted automatically. When adjustment is complete, select Done. 5 Select "Continue" to test the dark current and the water blank AD of each channel. 6 Select "Continue" to finish the alignment and exit the procedure. NOTE: Ensure that the screws on the reaction carousel position sensor bracket have been tightened. BS-240 Chemistry Analyzer Service Manual 8-9 8 Alignment 8.3 Photoelectric Unit Alignment parameters can be saved only after each procedure is complete. Another alignment is required if a procedure is terminated. If the parameter is not successfully configured or out of range, the parameter range will be displayed by the alarm message and re-perform the alignment. 8-10 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.4 Sample/reagent carousel unit 8.4 Sample/reagent carousel unit Select Utility -> Maintenance -> Alignment -> Sample/Rgt Carousel. The window is shown below. 8.4.1 Reaction Carousel Rotary Alignment index: The sample probe (pseudo probe BA60-J07) can pass through the through-hole on the cuvette alignment fixture (BA24-J04). Operation methods and steps: 1 Select Reaction Carousel Rotary. Access the alignment screen and operate as prompted. 2 Select "Continue" to reset all mechanical parts of the whole unit. 3 Place a cuvette alignment fixture (BA24-J04) in cuvettes of the sample position 19# and dilution position 14# of the reaction carousel body 4 Adjust the sample probe to the position of the reaction carousel. Tighten the screw of the arm of the sample probe to such an extent that the arm cannot swing freely but can rotate freely when you push the arm. Note: If the arm can rotate automatically, the position is not proper. When you push the arm manually, do not drive rotation of the motor. If the motor rotates, exit the window and access the alignment window again. BS-240 Chemistry Analyzer Service Manual 8-11 8 Alignment 8.4 Sample/reagent carousel unit Click the button in the dialog box, rotate the probe to the top of the sample dispense position of the reaction carousel, and then use a hand to push the probe arm to swing it, and observe whether the probe can pass through the center hole of the alignment fixture on two cuvette positions simultaneously. If it cannot, adjust the radial position of the probe and the circumferential parameters of the probe (the left and right arrows on the lower right corner of the screen control the circumferential movement of the carousel) to achieve the alignment purpose. After the radial position of the probe is confirmed, tighten the radial retaining screw. BA60-J07 BA24-J04 Circumferential track of reaction carousel Probe track Find an intersection at reaction cuvette center between the probe track and the circumferential track of reaction carousel After completion of the alignment, select "Continue" to proceed to the next step. 5 Click "Continue" to check whether the positions of the dispense position and the dilution position of the reaction carousel are correct. 6 Select "Continue". The alignment is complete. Select Exit to close the window. NOTE: The alignment fixtures will be used in following alignment. Leave them aside. The left arrow button is equivalent to rotating clockwise and the right arrow button rotating counterclockwise. When selecting Large Step, you can set up the concrete steps according to the reference range. 8-12 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.4 Sample/reagent carousel unit Once the probe arm length is determined, it must not be adjusted during alignment of probes and carousels. If the probe arm length is changed, the probe and carousel units should be aligned again. Alignment parameters can be saved only after each procedure is complete. Another alignment is required if a procedure is terminated. If the parameter is not successfully configured or out of range, the parameter range will be displayed by the alarm message and re-perform the alignment. 8.4.2 Sample/Reagent Carousel Outer Ring Radial Position Alignment index: The sample probe (fixture: BA60-J07 pseudo probe) can pass through the center hole of the sample position alignment fixture (BA48-J10) in sample positions 1#, 14#, and 27# of the outer ring of the sample/reagent carousel. Alignment methods and procedure: 1 Select "Sample/Reagent Carousel Inner Ring Circumferential Position", and then operate according to the screen prompts. 2 Place a sample position alignment fixture BA48-J10 in sample positions 1#, 14#, and 27# of the outer ring of the sample/reagent carousel. Select "Continue". The probe tip can pass through the central hole of the alignment tool. BA48-J10 3 Loosen the arm screw a little so that the probe arm cannot rotate freely but can be rotated by hands. If the arm rotates freely, the adjusted position will be inaccurate. Do not move the motor when rotating the probe arm. If the motor moves, exit the window and access again. BS-240 Chemistry Analyzer Service Manual 8-13 8 Alignment 8.4 Sample/reagent carousel unit 4 Manually rotate the probe arm to 1# position of the sample/reagent carousel outer ring, and click the up and down arrow buttons to stop the probe at the visually nearest position above the fixture. Check if the probe tip is aligned with the middle of the through-hole. If it is not, manually rotate the arm to the crossing point in circumferential direction of the sample carousel outer ring, and click the left and right arrow buttons to adjust the probe's circumferential position on the outer ring, so that the probe tip can pass through the center hole of the fixture in 1# position. The 4 arrows are used to control the probe. The 2 arrows are used to control the carousel. 5 Adjust the reagent probe in 14# and 27# positions of the sample/reagent carousel outer ring according to step 3. 6 Select "Continue". The alignment is complete. Select Exit to close the window. 8.4.3 Sample/Reagent Carousel Middle Ring Radial Position Alignment index: The sample probe (fixture: BA60-J07 pseudo probe) can pass through the center hole of the reagent center alignment fixture (BA10-J16) in reagent positions 1#, 14#, and 27# of the middle ring of the sample/reagent carousel. Alignment methods and procedure: 1 8-14 Select "Sample/Reagent Carousel Middle Ring Radial Position". BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.4 Sample/reagent carousel unit 2 Place a reagent position alignment fixture BA10-J16 in positions #1, #14 and #27 of the sample/reagent carousel, and then select "Continue". 3 Adjust the positions #1, #14 and #27 of the middle ring of the sample/reagent carousel according to step 4 of "Sample/Reagent Carousel Inner Ring Circumferential Position". 4 Select "Continue". The alignment is complete. Select Exit to close the window. 8.4.4 Sample/Reagent Carousel Inner Ring Radial Position Alignment index: The sample probe (fixture: BA60-J07 pseudo probe) can pass through the center hole of the reagent center alignment fixture (BA10-J16) in reagent positions 41#, 54#, and 67# of the inner ring of the sample/reagent carousel. Alignment methods and procedure: 1 Select "Sample/Reagent Carousel Inner Ring Radial Position". 2 Place a sample position alignment fixture BA10-J16 in positions 41#, 54#, and 67# of the sample carousel, and then select "Continue". 3 Adjust the positions 41#, 54#, and 67# of the inner ring of the sample/reagent carousel according to step 4 of "Sample/Reagent Carousel Inner Ring Circumferential Position". 4 Select "Continue". The alignment is complete. Select Exit to close the window. NOTE: When the alignment is complete, there is no need to take out the alignment fixture. Alignment parameters can be saved only after each procedure is complete. Another alignment is required if a procedure is terminated. If the parameter is not successfully configured or out of range, the parameter range will be displayed by the alarm message and re-perform the alignment. BS-240 Chemistry Analyzer Service Manual 8-15 8 Alignment 8.5 Probe unit 8.5 Probe unit Select Utility—>Maintenance—>Alignment—>Sample Probe Unit. The window is shown below. 8.5.1 Probe to Horizontal Position on Reaction Carousel Alignment index: The sample probe (pseudo probe BA60-J07) can pass through the through-hole on the cuvette alignment fixture (BA24-J04). Alignment methods and procedure: 1 Select Sample Probe Rotary to Cuvette. 2 Place the cuvette alignment fixture (BA24-J04) in positions 19# and 14# of the reaction carousel, and then select "Continue". 3 Select the down arrow button to lower the sample probe to the top of the alignment fixture in position 19# so that it can align with the center hole. 4 Adjust the circumferential position of the probe arm and ensure that the sample probe can enter the center hole on the fixture. Press the probe arm and tighten the screws on it. If error exists, click the left and right arrow buttons to adjust the probe till it meets the requirement. 8-16 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.5 Probe unit BA24-J04 5 Select "Continue" and verify the alignment with the up and down arrow buttons. If the index is not satisfied, return to the previous step and try again. 6 Click "Continue". Move the sample probe to the top of the position 14#. Click the up and down arrows to drop the sample probe to the top of the fixture. Check whether the probe tip can enter the center hole of the fixture. If it cannot, click the left and right arrows to adjust parameters until index requirements are met. 7 Select "Continue" and verify the alignment with the up and down arrow buttons. If the index is not satisfied, return to the previous step and try again. 8 Select "Continue" to save parameters and exit the window. 8.5.2 Aligning Horizontal and Vertical Positions from Probe to Wash Well Alignment index: Horizontal position: The sample probe tip can enter the center hole on the wash well alignment fixture (BA48-J12). The gap between the wash well and the cover is even. Vertical position: When reaching the bottom of the center hole on the wash well alignment fixture (BA48-J12), the sample probe is lifted for 0.15mm, which means that the clearance gauge can pass with 0.15mm through the gap between the sensor plate and the arm base and it cannot pass with 0.20mm. Alignment methods and procedure: 1 Select "Sample Probe Rotary to Wash Well and Home". 2 According to step 2 on the screen, place a wash well alignment fixture (BA48-J12) in the sample probe wash well, and then select "Continue". 3 Select the up/down and left/right arrow buttons to move the sample probe till it lowers into the center on the alignment fixture. If the reagent probe is greatly deviating from the alignment fixture, use a big cross head screwdriver to loosen the screws that fasten the wash well or the wash well bracket, adjust the wash well till the sample probe can enter to the center hole on the alignment fixture, and then tighten the screws that fasten the wash well (and the wash well bracket). BS-240 Chemistry Analyzer Service Manual 8-17 8 Alignment 8.5 Probe unit Use the clearance gauge to measure the gap here and keep it away from the slit on the photocoupler side. The soldering lug is deformable and may affect the measurement. Criteria: The clearance gauge can pass with 0.15mm and cannot pass with 0.2mm. BA48-J12 4 Confirm the alignment result. If the requirements are not met, return to the previous step. 5 Adjust the sample probe's vertical position above the wash well by using the up and down arrow buttons. Check that the clearance gauge can pass with 0.15mm and cannot pass with 0.20mm. 6 Check the vertical position. If the requirements are not met, return to step 5. 7 Remove the alignment fixture according to screen prompts. 8 Select "Continue" to save parameters and exit the window. 8.5.3 Probe to Sample/Reagent Carousel Outer Ring Alignment index: The sample probe tip can enter the center hole of the alignment fixture (BA48-J10) in positions 1#, 14#, and 27# of the outer ring of the sample/reagent carousel. Alignment methods and procedure: 1 Select Sample Probe Rotary to Sample/Reagent Carousel Outer Ring. 2 Place a sample position alignment fixture (BA48-J10) in positions 1#, 14#, and 27# of the outer ring of the sample/reagent carousel, and then Select "Continue". 3 Select the up/down and left/right arrow buttons to move the sample probe so that it can enter into the center hole on the alignment fixture. 4 Select "Continue" to save parameters and exit the window. 8.5.4 Probe to Sample/Reagent Carousel Middle Ring Alignment index: The sample probe tip can enter the center hole of the alignment fixture (BA10-J16) in positions 1#, 14#, and 27# of the middle ring of the sample/reagent carousel. Alignment methods and procedure: 1 8-18 Select "Sample Probe Rotary to Sample/Reagent Carousel Middle Ring". BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.5 Probe unit 2 Place a reagent position alignment fixture (BA10-J16) in positions 1#, 14#, and 27# of the middle ring of the sample/reagent carousel, and then Select "Continue". 3 Select the up/down and left/right arrow buttons to move the sample probe so that it can enter into the center hole on the alignment fixture. 4 Select "Continue" to save parameters and exit the window. 8.5.5 Probe to Sample/Reagent Carousel Inner Ring Alignment index: The sample probe tip can enter the center hole of the alignment fixture (BA10-J16) in positions 41#, 54#, and 67# of the inner ring of the sample/reagent carousel. Alignment methods and procedure: 1 Select "Sample Probe Rotary to Sample/Reagent Carousel Inner Ring". 2 Place a reagent position alignment fixture (BA10-J16) in positions 41#, 54#, and 67# of the inner ring of the sample/reagent carousel, and then Select "Continue". 3 Select the up/down and left/right arrow buttons to move the sample probe so that it can enter into the center hole on the alignment fixture. 4 Select "Continue" to save parameters and exit the window. 8.5.6 Sample Probe Cuvette Height Alignment index: When reaching the cuvette bottom, the sample probe is lifted for 0.15mm, which means that the clearance gauge can pass with 0.15mm through the gap between the sensor plate and the arm base and it cannot pass with 0.20mm. Alignment methods and procedure: 1 Select "Adjust Vertical Extreme Position from Probe to Reaction Carousel". 2 Check that cuvettes are loaded in positions 1#, 14#, and 27# of the reaction carousel. 3 Adjust the sample probe over the 3 cuvette positions, and ensure that the lifting height meets the requirement. 4 Select "Continue" to save parameters and exit the window. 8.5.7 Vertical Extreme Position from Probe to Reagent Bottle Alignment index: When reaching the reagent bottle bottom, the sample probe is lifted for 0.15mm, which means that the clearance gauge can pass with 0.15mm through the gap between the sensor plate and the arm base and it cannot pass with 0.20mm. BS-240 Chemistry Analyzer Service Manual 8-19 8 Alignment 8.5 Probe unit Alignment methods and procedure: 1 Select "Adjust Vertical Extreme Position from Probe to Reagent Bottle". 2 Check that reagent bottles are loaded in positions 1#, 14#, and 27# of the middle ring of the sample/reagent carousel. 3 Adjust the sample probe over the 3 reagent positions, and ensure that the lifting height meets the requirement. 4 Select "Continue" to save parameters and exit the window. 8.5.8 Sample Probe Rotary to ISE (Optional) Alignment index: The sample probe has its tip above the ISE sample injection port but deviates slightly from the center hole. The probe tip should not contact with the inwall of the sample inject port. Alignment methods and procedure: 1 Check if the ISE module has been installed. 2 Select "S Probe Rotary to ISE". 3 According to step 3, select the down arrow button to lower the sample probe to above the ISE sample injection port. Select the left/right arrow buttons to adjust the sample probe radially till it is 1mm away from the center of the sample injection port or deviating slightly from the center hole. The probe tip should not contact with the inwall of sample inject port. 4 If the sample probe fails by adjusting parameters, loose the screws fixing the ISE module on the chassis, and then move it till requests are met. 5 If the problem still remains, use a cross head screwdriver to loosen the screws on the ISE module and move it till requests are met. 6 The operation in step 3 is verified in step 4. If requirements are not met, return to step 3. 7 Select "Continue" to save parameters and exit the window. 8.5.9 Sample Probe Rotary to SIC Height (Optional) Alignment index: The sample probe has its tip level with the third transition line in the sample injection cup. Alignment methods and procedure: 8-20 1 Check if the S Probe Rotary to ISE alignment has been completed. 2 Select "S Probe Rotary to SIC Height". BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.5 Probe unit 3 According to step 3 on the screen, select the up/down arrow buttons to make the probe tip level with the third transition line inside the sample injection cup. (You can see three transition lines, and the probe tip shall be level with the one at the bottom, as shown in the figure below.) 4 The operation in step 3 is verified in step 4. If requirements are not met, return to step 3. 5 Select "Continue" to save parameters and exit the window. Probe tip level with this line NOTE: Alignment of each position is verified the next step. Alignment parameters can be saved only after each procedure is complete. Another alignment is required if a procedure is terminated. If the parameter is not successfully configured or out of range, the parameter range will be displayed by the alarm message and re-perform the alignment. BS-240 Chemistry Analyzer Service Manual 8-21 8 Alignment 8.6 Mixer Unit 8.6 Mixer Unit Select "Utility—>Maintenance—>Alignment—>Mixer Unit". The window is shown below. 8.6.1 Adjusting Horizontal Position from Mixer to Reaction Carousel Alignment index: The mixer (using BA48-J16 mixer alignment lever as a fixture) can enter the through hole of the BA24-J05 mixing horizontal position alignment cuvette in position 9# of the reaction carousel. Alignment methods and procedure: 1 8-22 Install the mixer fixture BA48-J16, and then select "Adjusting Horizontal Position from Mixer to Reaction Carousel". BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.6 Mixer Unit BA48-J16 : Alignment lever (inserted to the end) 2 Place 1 mixer-reaction cuvette horizontality alignment fixture (BA24-J05) in position 9# of the reaction carousel. 3 Adjust the mixer position according to step 3. Click the up and down arrow to move the mixer to the top of the fixture. When the mixer can be aligned with the center position of the fixture on the circumferential track, use an M3 hexagon wrench to pre-tighten the arm screw (to such an extent that the arm cannot swing freely but can rotate freely when you push the arm manually). Rotate the arm to a lower position that has enough space. Pre-tighten the radial screw of the mixer, and then rotate back to the top of the reaction carousel fixture. Determine the radial position of the mixer, and then rotate the arm to a position that has enough space, and tighten the radial screw. The radial retaining screw of the mixer is on the lower part, and the retaining screw space is deficient, so that the arm needs to be moved to sufficient space to facilitate operation BA48-J16 BA24-J05 Return to the top of the reaction carousel fixture and fasten the circumferential screw of the arm. 4 If the lever is not correct horizontally, select the left/right arrow buttons in step 4 to adjust so that it can align with the alignment fixture. 5 The operation in step 4 is verified in step 5. If requirements are not met, return to step 4. 6 Select "Continue", remove the fixture and store it properly. BS-240 Chemistry Analyzer Service Manual 8-23 8 Alignment 8.6 Mixer Unit 8.6.2 Adjusting Vertical Position from Mixer to Reaction Carousel Alignment index: The lever drops to the center of the cuvette, so that the go gauge of the block of the alignment fixture (BA80-J20-001) can pass through the gap between them and the no-go gauge cannot. Alignment methods and procedure: 1 Select "Adjust Vertical Position from Mixer to Reaction Carousel". Make sure that the fixture of the mixing position is holding a cuvette instead of the alignment fixture. 2 Select "Continue". After the system resets mechanically, the mixer moves into the cuvette. 3 Select the up/down arrow buttons, and insert the alignment fixture BA80-J20-001 between the mixer's straight-knurled column and cuvette edge. Gently push the alignment fixture to observe if the Go Gauge can easily pass while the No-go Gauge cannot. The go gauge of mixer alignment tool (thin) can be passed through and no-go gauge (thick) cannot be passed through 4 The operation in step 2 is verified in step 3. If requirements are not met, return to step 2. 5 Select "Continue" to save parameters and exit the window. 8.6.3 Adjusting Horizontal Position from Mixer to Wash Well Alignment index: The mixer (using BA48-J16 mixing alignment lever as a fixture) can enter the through hole on the alignment fixture (BA40-J76) of the position from the mixer to the wash well, and the gap between the mixer/wash well and the cover is even. Alignment methods and procedure: 1 8-24 Select "Adjust Horizontal Position from Mixer to Wash Well". BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.6 Mixer Unit 2 Click "Continue". Place a wash well alignment fixture BA40-J76 in the mixing wash position. 3 Click the up and down arrow to move the mixer fixture to the top of the wash well alignment fixture. Click the left and right arrows so that the circumferential position of the mixer fixture can be aligned with the center hole of the wash well alignment fixture. If the radial position cannot be aligned, loosen the wash well retaining screw, and adjust the position of the wash well to meet alignment requirements. BA40-J76 4 The operation in step 2 is verified in step 3. If requirements are not met, return to step 2. 5 Remove the fixture, Select "Continue" to save parameters and exit the window, and then replace the mixer. BS-240 Chemistry Analyzer Service Manual 8-25 8 Alignment 8.7 Setting Parameters of Cuvette Wash Unit 8.7 Setting Parameters of Cuvette Wash Unit Select Utility—>Maintenance—>Alignment—>Cuvette Wash Unit. The window is shown below. 8.7.1 Setting Parameters of Best Alignment Position Alignment methods and procedure: 1 Search the H-046-008697-00 BS-240 Series Parameter Configuration Table for the best alignment cuvette segment number. Input the best cuvette segment number of the cuvette wash unit, and the maximum deviation cuvette segment in positive and negative directions, and then click "Configure". Select "Search" and check whether the configuration result is correct. 8-26 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.7 Setting Parameters of Cuvette Wash Unit 8.7.2 Adjust Horizontal Position from Wash Probe to Reaction Cuvette Alignment index: The wash probe can pass through the hole of the wash probe alignment fixture (BA24J01), and move smooth. Alignment methods and procedure: 1 Select "Adjust Horizontal Position from Wash Probe to Reaction Cuvette". 2 Remove the reaction cuvette segment in the best cuvette segment position, place a fixture BA24-J01 in the position, and Select "Continue". 3 According to Step 3 on the screen, use an M4 hexagon wrench to loosen the two screws on the wash probe conversion board, adjust the wash probes till the wipe blocks can move vertically in the center hole of the alignment fixture. When requests are met, tighten the two screws gradually. Exercise caution to prevent deforming the probes. BS-240 Chemistry Analyzer Service Manual 8-27 8 Alignment 8.7 Setting Parameters of Cuvette Wash Unit BA24-J10 4 Remove the fixture, change back the reaction cuvette segment, click the up and down arrows, and control the wash probe to move vertically in the reaction cuvette. The movement should be smooth without interfering with the cuvette wall. If the requirements are not met, return to step 3 and adjust again. 5 In Step 5, the reaction carousel rotates to the forward maximum deviating cuvette segment. Click the up and down arrows to move the wash station so that the wipe blocks and wash probes are in the center of the cuvettes and can move smoothly. 6 In Step 6, the reaction carousel rotates to the backward maximum deviating cuvette segment. Click the up and down arrows to move the wash station so that the wipe blocks and wash probes are in the center of the cuvettes and can move smoothly. 7 Select "Continue" to save parameters and exit the window. 8.7.3 Cuvette Wash Station Height Alignment index: When the wash station lowers to the vertical washing position, the wash probes bounce by 0.7 to 0.8mm, which means that the clearance gauge can pass with 0.7mm through the gap between the anti-collision blocks and the wash station bracket and it cannot pass with 0.8mm. Alignment methods and procedure: 8-28 1 Select "Adjust Vertical Position from Wash Probe to Reaction Cuvette". 2 Check if the wipe blocks can insert into the center of each cuvette. Select "Continue". BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.7 Setting Parameters of Cuvette Wash Unit 3 When the wash probes move vertically to the bottom of the cuvettes, make the phase-4 probe contact the cuvette bottom, and select Small Step or Large Step. Adjust the probes with the up/down arrow buttons while measuring the lifting height with 0.7mm and 0.8mm clearance gauges. Make sure the gap between the anti-collision blocks and the wash station bracket is enough for passing through the 0.7mm clearance gauge but not enough for passing through the 0.8mm clearance gauge. 4 Verify Step 3. If requests are not met, return to Step 3, and select the up/down arrow buttons to adjust the mixer till requests are met. 5 Select "Continue" to save parameters and exit the window. BS-240 Chemistry Analyzer Service Manual 8-29 8 Alignment 8.8 Bar Code Unit (Optional) 8.8 Bar Code Unit (Optional) 8.8.1 Bar Code Sticking Requirements 1 The bar code label should be clear, and not vague, dirty or scratched. 2 The sample bar code label should be stuck upright and plane and 9mm away from the tube bottom. Sample tube Area for affixing bar code 9mm 3 The reagent bar code label should be stuck in the front side below the bottle opening, and in the middle of the vertical and horizontal directions. Do not stick to the corner 8-30 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.8 Bar Code Unit (Optional) Do not stick to the corner 8.8.2 Bar code Unit Select Utility -> Maintenance -> Alignment -> Bar code. The window is shown below. 8.8.3 Adjusting Bar Code Reader Position Alignment index: BS-240 Chemistry Analyzer Service Manual 8-31 8 Alignment 8.8 Bar Code Unit (Optional) The light beam transmitted from the reagent bar code reader can pass through the slit in the sample position on the bar code reader fixture (BA24-J03) and the slit in the reagent position. The bar code in outer ring positions 1~5# and middle ring positions 1~5# of the sample/reagent carousel can be identified correctly and completely and displayed on the screen. Alignment methods and procedure: 1 Select "Adjust Bar Code Reader". 2 Use a cross screwdriver to remove the tube holder base in the outer positions 26# and 27#, place the bar code alignment fixture BA24-J03 (the notch surface of the fixture faces outside), and Select "Continue". Remove the sample rack, and insert the lower positioning pin of BA24-J03 into the two holes 3 Reagent bar scan gap code Sample bar scan gap code Loosen 4 screws that fasten the bar code reader and the bracket. Adjust the angle of the light beam until the bar code reader light can pass through the slit in the sample position of the fixture. If a small circumferential deviation exists, click the left and right arrows on the screen, and adjust the sample/reagent carousel parameters to meet requirements. 8-32 4 Check again if the light goes through the slot on the fixture. If not, return to previous step to make adjustment. 5 Remove the alignment fixture, load bar-coded sample tubes in positions 1#-5# of outer ring and 1#-5# of middle ring, and then start scanning. 6 If all bar code cannot be read, check if the light direction meets the requirement and the bar code reader works normally. 7 Select "Continue" to save parameters and exit the window. BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.8 Bar Code Unit (Optional) Place white paper in front of the fixture, and the alignment succeeds when the scan light penetrates the gap on the fixture and falls on the paper to form a complete straight line Adjust the screw here (with two more screws below), and adjust the location of the bar code reader 8.8.4 Bar Code Stability Test Alignment index: When the sample/reagent carousel rotates for several circles, the bar code identified in the same position should be the same. The bar code in outer ring positions 1~40# and middle ring positions 1~40# of the sample/reagent carousel can be identified correctly and completely and displayed on the screen. Randomly select 3 bar codes on the outer ring and the middle ring separately, and the software display is consistent with the affixed bar code. Alignment methods and procedure: 1 Place bar-coded tubes and reagent bottles in sample positions #1- 40# on the outer ring of the sample carousel, and in reagent positions #1- 40# on the middle ring of the reagent carousel. 2 Select Reagent Bar Code Stability Test, set the scan circle as 10, and select "Start". 3 After scanning, check if the identified bar code meets the requirement. If different bar code is read in the same position, the position will be indicated in red. 4 If the stability test fails, find the reasons and take actions according to the following probabilities: Check if the failed bar code meets the printing standard, the bar code label is not tilting, dirty or damaged and has been applied in correct position. Check if the glass scanning window is clean without contamination. The scanning window is not blocked by the sponge cushion. If none of the above happens, replace the bar code reader, adjust it properly and do the test again. Generally, the three methods above are enough to troubleshoot the problem. NOTE: Check that the sponge cushion seals the bar code reader completely without blocking the scanning light. Otherwise, adjust the bar code reader again. BS-240 Chemistry Analyzer Service Manual 8-33 8 Alignment 8.8 Bar Code Unit (Optional) After alignment, check if the screws on the bar code reader and conversion board have been tightened. 8-34 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.9 Hydropneumatic Unit 8.9 Hydropneumatic Unit Select Utility—>Maintenance—>Alignment—>Hydro Unit. The window is shown below. 8.9.1 Preparations before Alignment 1 Check if all tubes have been connected correctly. 2 Check if the transit tubes on the fourth phase wash probe assembly have been connected properly. Visually check that the fourth phase aspirate tubes are inserted with the same depth and not bent, and all tubes are smooth without crossover or protrusion. 3 Check if the solenoid valves have connected to the correct cables. 4 Enough diluted wash solution exists in the diluted wash solution bottle. The dilution factor of the wash solution is configured according to the cuvette type. Cuvette type Plastic Cuvette Wash Solution Dilution Factor 1 8~12 (water):1 (concentrated wash solution) 5 Fill the water tank with water. 6 Connect the water supply tubing to the instrument, and connect the high-/lowconcentration waste tanks. 7 Check if the inlet filter assembly has been installed. 8 Remove the pseudo probes from the sample probe arm, and install the real probes instead. BS-240 Chemistry Analyzer Service Manual 8-35 8 Alignment 8.9 Hydropneumatic Unit Installation requirements: (1) Gap between the probe plate and the arm surface is not allowed; otherwise, check the retaining bolt and tube, because the tube's elasticity will also affect the probe's bouncing; (2) a washer has been applied on the retaining bolt, and the nut has been tightened; (3) after fixing the probe with bolt and spring, manually lift the probe and release it, and check that the probe can bounce back freely through the spring; (4) Adjust the level sense board and the relative positions of the probe plate and photocoupler so that the probe plate is in the middle of the photocoupler. 8.9.2 Checking Floater and Valve Status Manually switch the floater status one by one according to the table below, and then select Query to refresh the floater status. Check if the actual fluid level of each container is the same as the status displayed on the screen. Tank Name Floater Status Screen Display Water tank, solution tank diluted wash Floater triggered (high level) Not empty Water tank, solution tank diluted wash Floater level) Empty not triggered (low High-concentration waste tank Floater triggered (high level) Full High-concentration waste tank Floater level) Not Full not triggered (low 8.9.3 Syringe Check 1 8-36 Select Syringe Check. BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.9 Hydropneumatic Unit 2 Click "deionized water syringe" and "wash solution syringe". If no alarm appears, the syringes are OK. 8.9.4 Fluidic Prime 1 Select "Fluidic Prime". 2 According to instruments on the window, make sure that the inlet and outlet tubing has been connected. Remove the wash probe, and place it in an empty container. 3 Enter step 3, click "Continue", and click "Start" to prime the wash probe. When phase 1-3 wash probes spray water continuously, click "Stop" to stop priming. Click "OK" to enter the next step. If the priming is not finished in the specified number of times, set a new number of times, and perform priming again. 4 Reinstall the wash station. According to the prompt on the window, execute the wash instruction twice, and confirm that no overflow occurs. 5 Click "Continue". According to the prompt on the software, perform mixer exterior washing and priming. When the mixer wash well can spray water continuously, click "Stop" to stop priming. Select "OK" to move to the next step. BS-240 Chemistry Analyzer Service Manual 8-37 8 Alignment 8.9 Hydropneumatic Unit 6 Click "Continue". According to the prompt on the software, perform probe interior and exterior washing and priming. When the probe interior and exterior are washed and water is sprayed continuously, click "Stop" to stop priming. Select "OK" to move to the next step. Note: When the software is run, remove the sample syringe of the probe, push and pull the syringe plunger repeatedly until air bubbles are removed completely from syringes and T pieces. After all air is discharged, click "Stop". Reinstall the syringe in such a way that the syringe scale 7.5 is aligned with the V notch. Click "OK" in the dialog box, and enter the next step. 7 Prime the phases 3-6 wash probes. Select "Continue" to pop up a countdown dialog box. The instrument primes the phase 3-6 wash probes. Check the dispense probes till those of phase 3-6 spray clear water continuously without bubbles. After completion of the priming, observe if the phase 3-6 wash probes are stopped in order. If yes, select OK; otherwise, select OK and then select Stop at the lower-right corner. Troubleshoot the error and prime the probes again. 8 Select "Continue" to reset the mechanical parts for 3 times. Check that the mechanical parts are moving normally. 9 Select "Continue". 10 Select Exit to finish the alignment procedure. Select Maintenance -> Parameters -> Wash Station to check if the value of "Fluidic Prime Complete" is 1. If yes, the prime is successful. 8-38 BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.10 Pyrology Unit 8.10 Pyrology Unit Select Utility -> Maintenance -> Alignment -> Pyrology Unit. The window is shown below. 8.10.1 Sensor Configuration window Alignment index: Configure the parameters correctly according to the rated values of the reaction carousel sensors, and configure the △AD value of temperature sensors. Alignment methods and procedure: 1 Select Sensor Configuration. 2 According to the sequence number on the temperature sensor, contact the headquarters to find the resistance of rated temperature points (0, 37 and 100), select sensor 1 and input the resistance for each temperature point, and then select Calculate. 3 When a dialog box pops up indicating qualified sensor, select Configure to configure the parameters to the relevant sensor. 4 Observe the △AD value of PCB, and input it in the edit boxes of the window for sensor 1. Select Configure, and then select Search to verify them. 5 When configuration is complete, select Exit. BS-240 Chemistry Analyzer Service Manual 8-39 8 Alignment 8.10 Pyrology Unit 8.10.2 Observe Temperature Curve Alignment index: The reaction carousel heating temperature is 36.7~37.3°C, and the reagent preheating temperature is 43~47°C. Alignment methods and procedure: 1 Select Reaction Carousel Temperature Curve. 2 Select "Turn on heater". 3 Select "Sensor Temp Curve 1"(namely, reaction carousel heating temperature curve). In the temperature curve query area, select "Start". After the curve is steady (when the temperature rises to about 37°C), observe the reaction carousel temperature curve for about 1 minute (which may be counted according to the time displayed on the computer). Make sure that the temperature curve of the sensor is straight and stable and fluctuate within 36.7~37.3°C. 8-40 4 Similarly, check whether the reagent preheating curve is in the specified range ("sensor temperature curve 2" is the reagent preheating curve). 5 Select "Stop". Select "Exit" to close the window. BS-240 Chemistry Analyzer Service Manual 8 Alignment 8.10 Pyrology Unit 8.10.3 Wash Solution Temperature Control Alignment index: The temperature of cleaning fluid and wash solution is within 39°C-49°C. Alignment methods and procedure: 1 Select Wash Station on the Parameter Configuration window, and check that the value of "Fluidic Prime Complete" is 1. Select Utility -> Maintenance -> Alignment -> Pyrology Unit -> Wash Solution Temp Curve. 2 Select Deionized Water Heater and Wash Solution Heater, and then select "Turn On". 3 Select "Start". Observe the temperature curves of deionized water and wash solution, and check if they are straight and stable (for about 2 minutes, counted by the computer time) with average value within 39°C-49°C. 4 Select "Stop". Select "Exit" to close the window. BS-240 Chemistry Analyzer Service Manual 8-41 8 Alignment 8.10 Pyrology Unit 8.10.4 Reagent Refrigeration Alignment Methods and procedure: 1 8-42 Keep refrigerating the reagent compartment for 30 minutes, and touch the reagent compartment with a hand to check the refrigeration effect. BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance BS-240 Chemistry Analyzer Service Manual 9-1 9 Proactive Maintenance 9.1 Overview 9.1 Overview Proactive maintenance is a preventive measure taken by service personnel of Mindray or authorized by Mindray with the aim of eliminating hidden troubles to ensure system reliability and achieve best performance during operation. The proactive maintenance cycle for the chemistry analyzer is one year and for ISE module, half a year. The maintenance includes changing, cleaning or checking the parts as shown in the table below. Table 9.1 Maintenance items to be performed by service personnel Category Replacement items Cleaning items Check Item 9-2 Procedure When to do Replace Water Inlet Filter More than 1 year Replace Diluted Wash Solution Filter More than 1 year Replace Syringe Over 300,000 times Replace Calibrator Tube(ISE configured) More than 1 year Replace Pump Tube(ISE configured) More than half year Replace Probe Washer More than 1 year Clean dust screens When needed Clean the dust of the heatsink fan When needed Clean Wash Wells When needed Clean Cuvette Wash Station and Tubes When needed Clean ISE Injection Port (ISE Configured) When needed Clean deionized water Tank When needed Clean Diluted Wash Solution Tank When needed Clean Sample/Reagent Compartment When needed Clean Analyzer Panels When needed Special Wash When needed Clean Probe Exterior When needed Clean Mixers When needed Pump Calibration (ISE Configured) When needed Air Bubble configured) When needed Detector Calibration(ISE Cuvette Check When needed Photometer Check When needed Check Waste Pump When needed Check deionized water Syringe When needed Check Wash Solution Syringe When needed BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance 9.2 Maintenance Tools 9.2 Maintenance Tools Table 9.2 Maintenance Tools Too Name Applicable Maintenance Philips-head screwdriver φ3.3×100 Removing the system enclosure and the cooling fans Philips-head screwdriver φ4.7×100 Installing and removing the probes and lamp Slot-head screwdriver φ4.7×100 Removing probe and pipe hoop Round-head cleaning 0.25±0.01mm×125mm round tip tool, Unclogging the probes Hair brush Cleaning filter core and dust screens Gauze Wipe probe/mixer exterior sample/reagent compartment Cotton swabs Clean Wash Wells Suction cleaner Cleaning fans and dust screens Tweezers Removing/Installing washers Beaker Maintaining wash station Pipette Clean ISE sample injection port Ethanol Cleaning photometer, probes, mixers and wash station NaClO (0.5% sodium hypochlorite solution) Clean wash wells BS-240 Chemistry Analyzer Service Manual probes and and syringe 9-3 9 Proactive Maintenance 9.3 Maintenance Procedure 9.3 Maintenance Procedure 9.3.1 Execution order of maintenance items Among the proactive maintenance items of BS-240, some of them are performed with the main power switched off, while the other are performed through the maintenance window on the operating software, where maintenance logs are recorded. To improve the efficiency and shorten the field maintenance time, you are recommended to follow the procedure below: Table 9.3 Execution order of maintenance items No 9-4 Procedure Maintenance Tools Operations 1 Clean Analyzer Panels Ethanol and gauze Wipe 2 Clean dust screens Hair brush, cleaner suction Remove dust 3 Clean the dust of the heatsink fan Hair brush, cleaner suction Remove dust 4 Replace Water Inlet Filter Filter Replace 5 Replace Diluted Wash Solution Filter Filter Replace 6 Clean deionized water Tank / Wash 7 Clean Diluted Wash Solution Tank / Wash 8 Replace Syringe Syringe Replace 9 Replace Probe Washer Washer Replace 10 Replace Pump Tube(ISE configured) Pump Tube Replace 11 Replace Calibrator Tube(ISE configured) Calibrator Tube Replace 12 Clean Probe Exterior Ethanol and gauze Wipe 13 Clean Mixers Ethanol and gauze Wipe 14 Clean Wash Wells NaClO solution cotton swabs 15 Clean Cuvette Wash Station Ethanol and swab 16 Clean Sample/Reagent Compartment Ethanol, gauze,cotton swab Disassemble and wipe 17 Clean ISE Injection Port (ISE Configured) Ethanol and swab Wipe 18 Pump Calibration (ISE Configured) / Execute 19 Air Bubble configured) / Execute 20 Special Wash Concentrated solution 21 Cuvette Check / Execute 22 Photometer Check / Execute Detector Calibration(ISE and Wipe cotton Wipe cotton wash Execute BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance 9.3 Maintenance Procedure 23 Check Waste Pump / Execute 24 Check deionized water Syringe / Execute 25 Check Wash Solution Syringe / Execute 9.3.2 Steps Clean Analyzer Panels 1 Open the upper protective shield of the analyzer. 2 Use clean gauze moistened with ethanol to clean the analyzer panels and carousel covers. 3 Use wash solution to clean the monitor screen and keyboard. 4 Restore the protective shield. 5 After maintenance, open the logs and record related information. BS-240 Chemistry Analyzer Service Manual 9-5 9 Proactive Maintenance 9.3 Maintenance Procedure Cleaning dust screens 1 Switch off the analyzer power. 2 Remove the left cover and the rear panel of the analyzer. 3 Remove the dust screen. For the Disassembling Steps, refer to 5.2.4 Removing left/right dust screen. 4 Tidy and clean the dust screens. 5 Dry the dust screens in air. 6 Restore the dust screen. 7 Reinstall the left cover and the rear panel of the analyzer. 8 After maintenance, open the logs and record related information. Cleaning the dust of the heatsink fan 1 Refer to Section 5.1.8 in Chapter 5 for cleaning dust of the fans. 2 After maintenance, open the logs and record related information. Replacing Water Inlet Filter 1 Check the maintenance log and when the age of the inlet filter is more than 1 year, change it. 2 Refer to Section 7.5.3 in Chapter disassembling/assembling the filter. 3 After maintenance, open the logs and record related information. 7 Cuvette wash module for Replacing Diluted Wash Solution Filter 1 9-6 Check the maintenance log and when the age of the diluted wash filter is more than 1 year, change it. BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance 9.3 Maintenance Procedure 2 Refer to Section 7.5.3 in Chapter disassembling/assembling the filter. 3 After maintenance, open the logs and record related information. 7 Cuvette wash module for Cleaning deionized water Tank 1 Refer to Section 7.5.4 in Chapter 7. 2 After maintenance, open the logs and record related information. Cleaning Diluted Wash Solution Tank 1 Refer to Section 7.5.4 in Chapter 7. 2 After maintenance, open the logs and record related information. Replacing Syringe 1 Check the using count of the syringe on the status screen. When the sample syringe has been used for 100,000times, change it. 2 Prepare a new syringe or its plunger assembly and washer, put the plunger head in the deionized water beaker to remove air from the syringe, and then moisten the washer in the deionized water. 3 Place the analyzing unit power to the OFF position. 4 Loosen the front cover retaining screw of the syringe. 5 Loosen counterclockwise the four retaining screws on top of the syringe, and then remove the screws and the fixing blocks. 6 Loosen counterclockwise the retaining screw at the bottom of the syringe and then remove it. 7 Hold the T piece with one hand and the syringe connector with the other hand. Loosen the syringe counterclockwise and then remove the washer. 8 Loosen the plunger guide cap counterclockwise, hold the plunger head and pull it slightly to remove the plunger assembly from the syringe. 9 Insert the plunger head of the new plunger assembly into the bottom of the syringe, and then tighten the retaining screw to fix the plunger head. Tighten the plunger nut clockwise. 10 Soak the new syringe connector in the deionized water beak, gently pull the plunger head to aspirate half syringe of deionized water, and then push the plunger head to remove the air. 11 If no washer is found in the T piece, put the new washer in the T piece. Hold the T piece with one hand and the syringe connector with the other hand, and then screw the T piece clockwise. 12 Install the syringe on the bracket. BS-240 Chemistry Analyzer Service Manual 9-7 9 Proactive Maintenance 9.3 Maintenance Procedure 13 Install the fixing blocks and four retaining screws while having the retaining screws not tightened. 14 Align the plunger head to the retaining screw at the bottom of the syringe, and then tighten clockwise the retaining screw. 15 Pinch the plunger guide cap to adjust the syringe height. For the sample syringe, make the syringe head exceed the upper fixing block by 7.5 scales. 16 Tighten the four retaining screws on the fixing blocks. 17 After finishing replacement, switch on the analyzer power. 18 Perform the "Home maintenance" procedure. Check the new syringe for leak and bubbles. If leak occurs, check the syringe and the connector. 19 Fasten the front cover of the syringe 20 The procedure is completed. Write the maintenance log and input the maintenance information and clear the using count of the syringe. Replacing Probe Washer 1 Check the maintenance log. When the sample probe washer has been used for 1 year and change it. 2 Remove the probe and take out the probe washer. 3 Install a new washer and reinstall the sample probe. 4 Run the operating software, and select Utility-Maintenance-MaintenanceBiochemistry Maintenance and then select Probe interior clean and prime. 5 After maintenance, open the logs and record related information. Replacing Pump Tube (ISE configured) 1 Check the maintenance log. When Pump Tube has been used for half year, change it. 2 Refer to Section 5.10.4 in Chapter 5 for replacing Pump Tube. 3 After maintenance, open the logs and record related information. Replacing Calibrator Tube (ISE configured) 1 View the maintenance log. When the Calibrator Tube has been used for more than 1 year, change it. 2 Refer to Section 5.10.5 in Chapter 5 for Replace Calibrator Tube. 3 After maintenance, open the logs and record related information. Cleaning Probe Exterior 9-8 1 Place the analyzing unit power to the OFF position. 2 Lift gently the probe to the utmost height and move it to the position convenient for maintenance. BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance 9.3 Maintenance Procedure 3 Use gauze soaked with ethanol to gently wipe the probe exterior. Clean the probe tip until it becomes clear without stain. Use gauze moistened with deionized water to clear the ethanol on the probe. Do not pull the probe vertically to prevent probe damage. 4 After the procedure, Restore the upper protective shield of the analyzer. Power on the analyzing unit and select "Home". 5 After maintenance, open the logs and record related information. Cleaning Mixers 1 Place the analyzing unit power to the OFF position. 2 Lift gently the mixer to the utmost height and move it to the position convenient for maintenance. 3 Use gauze soaked with ethanol to gently wipe the mixer exterior. Clean it until it becomes clear without stain. Use gauze moistened with deionized water to clear the ethanol on the mixer. Do not pull the mixer vertically to prevent probe damage. 4 After the procedure, Restore the upper protective shield of the analyzer. Power on the analyzing unit and select Home. 5 After maintenance, open the logs and record related information. Cleaning Wash Wells 1 Place the analyzing unit power to the OFF position. 2 Move the probe and mixer away from the wash well for the convenience of maintenance. 3 Use clean cotton swabs moistened with NaClO solution to clean the wash wells. 4 After finishing replacement, switch on the analyzer power. 5 Execute "Home" command or Check Probes/Mixers/Wash Wells maintenance command. Check if the wash wells have a normal water flow. 6 After maintenance, open the logs and record related information. Cleaning Cuvette Wash Station 1 The analyzing unit should be powered off, or in failure mode or in standby or incubation status. 2 Open the upper protective shield of the analyzer. Remove the wash station, and use gauze moistened with ethanol to clean the nozzles and wipe blocks. 3 Use gauze moistened with deionized water to clear the ethanol on the wash probes. 4 Reinstall the wash station and close the upper shielding cover of the analyzer. 5 Execute the "Prime Wash Station" instruction on the "Maintenance" window. 6 After maintenance, open the logs and record related information. BS-240 Chemistry Analyzer Service Manual 9-9 9 Proactive Maintenance 9.3 Maintenance Procedure Cleaning Sample/Reagent Compartment 1 The analyzing unit should be powered off, or in failure mode or in standby or incubation status. 2 Remove the carousel cover and carousel, and then store them properly. 3 Use clean gauze soaked with deionized water or ethanol to clean the interior of the compartment . If necessary, you can use gauze moistened with neutral wash solution. 4 Use clean gauze soaked with deionized water or ethanol to clean the carousel body and bar code scanning window, and then use cotton swabs dipped with ethanol to clean the sample or reagent positions. 5 Reinstall the carousel into the compartment and restore the carousel cover. 6 After maintenance, open the logs and record related information. Cleaning ISE Injection Port (ISE Configured) 1 Make sure that the system status is Incubation or Standby. 2 Enter ISE maintenance screen and select "Clean ISE injection port". 3 Open the upper protective shield of the analyzer. 4 Open the cover of the ISE module on the analyzer’s front panel. 5 Use clean cotton swab soaked with ethanol to wipe the sample injection port (interior of the sample injection cup of the ISE module ) until it is clean; then use a clean cotton swab soaked with deionized water to wipe the interior and periphery of the sample injection port . 6 Enter the ISE Maintenance window and execute "purge A" and "purge B" each for three times. 7 Restore the cover of the ISE module on the analyzer’s front panel. Restore the upper protective shield of the analyzer. 8 After maintenance, open the logs and record related information. Pump Calibration (ISE Configured) 1 Make sure that the system status is Incubation or Standby. 2 Enter ISE maintenance screen and click Pump Calibration. 3 Fill a sample tube with at least 500μl deionized water, and then place it in No.40 position of the sample carousel. 4 Continue the procedure until it is completed. 5 After the procedure is completed, the calibration results are displayed. If calibration is successful, the procedure is completed. If not, ISE enters error status. 6 After maintenance, open the logs and record related information. Air Bubble Detector Calibration (ISE configured) 1 9-10 Ensure ISE is standby. BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance 9.3 Maintenance Procedure 2 Enter ISE maintenance screen, and select Air Bubble Detector Calibration. The procedure is performed automatically. 3 After the procedure is completed, the calibration results are displayed. If calibration is successful, the procedure is completed. If not, ISE enters error status. 4 After maintenance, open the logs and record related information. Special Wash 1 Open the upper protective shield of the analyzer. 2 Place more than 40 mL concentrated wash solution in position 39 of the reagent carousel. 3 Select Utility – Maintenance – Maintenance – Biochemistry Maintenance, and then select "Special Wash". 4 Confirm if cuvette check is needed after the concentrated wash. If it is, mark the checkbox in front of Check Cuvettes. 5 Select "Continue". The system starts cleaning the sample probe, mixers, cuvettes and wash station. When special wash is finished, the system performs cuvette check automatically. 6 Select "Done". 7 Restore the upper protective shield of the analyzer. 8 After maintenance, open the logs and record related information. Cuvette Check 1 Select Utility-Maintenance-Maintenance- Biochemistry Maintenance and then select "Cuvette check". 2 After check, check the cuvette check results for yellow indication. Record the cuvettes highlighted in yellow and wash or replace the cuvettes via a "Replace Cuvettes" procedure. 3 After maintenance, open the logs and record related information. Photometer Check 1 Select Utility-Maintenance-Maintenance- Biochemistry Maintenance and then select "Photometer Check". 2 If the result is Normal, the system will give no alarms, otherwise the following alarm will be given. If the alarm indicates the lamp is off, check if the lamp has been turn on. If the alarm light too strong is given, please perform photometric unit alignment or check if the photometric unit is normal. If the alarm indicates light intensity too weak, replace the lamp. BS-240 Chemistry Analyzer Service Manual 9-11 9 Proactive Maintenance 9.3 Maintenance Procedure Checking Waste Pump 1 Power off the analyzing unit, and remove the instrument cover and the upper cover assembly of the reaction carousel. Run the software to reset the system, and make the instrument become standby. Use a flat plate to cover the top of the photometer so that the dark current detection is performed upon startup. 2 Select Utility-Maintenance-Maintenance- Biochemistry Maintenance, and then select "Wash Cuvette" so that the system washes the reaction cuvette. 3 In the process of washing the reaction cuvette, check the liquid volume of the cuvette. If the liquid volume is flush with the cuvette top or overflows the cuvette top, remove the rear panel, and check and replace the problematic waste pump. The first phase probe of the wash station corresponds to P02, and the second phase probe and the third phase probe correspond to P03. If the fourth phase probe is unable to aspirate the leftover liquid completely, check and replace P04. 4 If the liquid volume is too low in the cuvette washing process, perform an "Wash and Prime" operation. 5 If everything is checked as normal, install the upper cover assembly on the reaction carousel and the instrument cover. 6 After maintenance, open the logs and record related information. Checking deionized water Syringe 1 Place the analyzing unit power to the OFF position, and remove the rear panel of the instrument. Run the software to reset the system, and make the instrument become standby. 2 Select Utility-Maintenance-Maintenance- Biochemistry Maintenance, and then select "Wash Cuvette" so that the system washes the reaction cuvette. 3 Check whether the 10ml wash syringe leaks. If it leaks, replace the wash syringe. For the Disassembling Steps, refer to Section 5.8.3 in Chapter 5. Open the maintenance logs and record related information. 4 If everything is checked as normal, reinstall the rear panel. Checking Wash Solution Syringe 1 Place the analyzing unit power to the OFF position, and remove the rear panel of the instrument. Run the software to reset the system, and make the instrument become standby. 2 Select Utility-Maintenance-Maintenance- Biochemistry Maintenance, and then select "Wash Cuvette" so that the system washes the reaction cuvette. 3 Check whether the 2.5ml wash syringe leaks. If it leaks, replace the wash syringe. For the Disassembling Steps, refer to Section 5.8.3 in Chapter 5. Open the maintenance logs and record related information. 4 9-12 If everything is checked as normal, install the upper cover assembly on the reaction carousel and the instrument cover. BS-240 Chemistry Analyzer Service Manual 9 Proactive Maintenance 9.3 Maintenance Procedure Checking Probe/Mixer Exterior Wash Flow 1 Open the upper cover of the analyzer and home the system. 2 Select Utility-Maintenance-Maintenance, Maintenance" command. 3 Choose "Prime Probe/Mixer Exterior". The maintenance guide window shows. Select "Continue". 4 Liquid flows out of the probe and mixer wash well automatically. Observe the liquid flow of the wash wells. Spillage, overflow or no liquid flow should not occur. The and then select "Biochemistry probe wash height should be no less than 5mm. The liquid in the mixer wash well should flow out of the its center and form a convex with a height of 1-2mm. 5 Repeat the above steps for three times to check if the probe/mixer flow is normal. If not, check if the exterior wash tubing is abnormal (Check first if the restrictor ring at the exterior wash valve outlet is clogged). Check the flow again. 6 Select "Done". 7 Close the maintenance window. BS-240 Chemistry Analyzer Service Manual 9-13 9 Proactive Maintenance 9.3 Maintenance Procedure 9-14 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting This chapter provides methods to locate and troubleshoot the errors and problems. Read this chapter thoroughly to achieve the best performance of the instrument. 10-1 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 9.3 Maintenance Procedure BS-240 Chemistry Analyzer Service Manual 10-2 10 Troubleshooting 10.1 Error Codes and Troubleshooting 10.1 Error Codes and Troubleshooting Table 10.1 Error Messages and Corrective Actions Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions A00006 Middle-layer unit Error Equipment configuration cannot be read or saved / E2PROM read/write error. Switch off the analyzing unit power and switch on it again. Recover failure by performing the Home maintenance procedure. If this message appears for 3 times, contact our customer service department or your local distributor. A00007 Middle-layer unit Error Instruction execute error. Specific error: Error details: / Instrument instructions cannot be executed. 1. Switch off the analyzer power and switch on it again. Recover failure by performing the Home maintenance procedure. 2. If this message appears for 3 times, contact our R&D department. A01006 Probe unit Error Probe movement Position: Error: Probe movement Position: Error: Probe movement Position: Error: / Probe vertical movement error 1. Sensor status error: Belt or belt pulley is loose.2. Failed to find the zero position: 1) The assembly is jammed; 2) Driving circuit error; 3) Vertical motor error; 4) Vertical sensor error; 5) Belt pulley is loose; 3. Collision occurs during operation other than aspirating: 1) Collision sensor or circuit or cable failure; 2) Collision sensor is not blocked when assembling the sample probe; 3) Incorrect sample probe positioning; 4) Incorrect sample/reagent carousel positioning; 4. Collision error Probe vertical movement error 1. Sensor status is incorrectCheck if belt or belt wheel is loose, tighten them if so; 2. Failed to find the zero position: 1) Remove obstacle or replace with a new assembly; 2) Replace PCB; 3) Replace motor; 4) Replace sensor; 5) Tighten properly; 3. Collision occurs during operation other than aspirating: 1) Replace sensor or PCB or cable; 2) Reassemble; 3) Rotate positioning sensor to remove dust or adjust probe to be vertical; if problem remains, adjust positioning parameters; 4) Adjust sample/reagent carousel positioning parameters; vertical error horizontal error syringe error (The specific type of error needs to be BS-240 Chemistry Analyzer Service Manual 10-3 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class BS-240 Chemistry Analyzer Service Manual Problem Description determined and displayed according to the movement result) Error: Probe vertical movement error 1. Sensor status is incorrect. 2. Failed to find the zero position. 3. Collision occurs during operation other than aspirating. 4. Collision error remains. 5. Moving vertically is not allowed in current horizontal position. Probe horizontal movement error 1. Sensor status is incorrect. 2. Failed to find the zero position. 3. Collision occurs during horizontal movement. 4. Moving horizontally is not allowed in current vertical position. Probe syringe movement error 1. Sensor status is Flag Probable Causes remains: 1) Collision sensor or circuit or cable failure; 2) Collision sensor is not blocked when assembling the sample probe; 3) Collision sensor is not blocked due to deformation of blocking plate’s driving spring; 5. Vertical movement is not allowed in current horizontal position: Software error Probe horizontal movement error 1. Sensor status error: Belt or belt pulley is loose.2. Failed to find the zero position: 1) Driving circuit error; 2) Motor error; 3) Home position sensor error; 4) Belt pulley is loose; 3. Collision occurs during horizontal movement: 1) Driving circuit error; 2) Motor error; 3) Home position sensor error; 4) Belt pulley is loose; 4. Horizontal movement is not allowed in current vertical position: Software error Probe syringe movement error 1. Sensor status error: Belt or belt pulley is loose.2. Failed to find the zero position: 1) The assembly is jammed; 2) Driving circuit error; 3) Motor error 4) Sensor error; 5) Belt pulley is loose. Corrective Actions 4. Collision error remains: 1) Replace sensor or PCB or cable; 2) Reassemble; 3) Mount the spring of blocking plate again; 5. Vertical movement is not allowed in current horizontal position:Home instrument or sample probe unit. Probe horizontal movement error 1. Sensor status is incorrectCheck if belt or belt wheel is loose, tighten them if so; 2. Failed to find the home position. 1) Replace PCB; 2) Replace motor; 3) Replace sensor; 4) Tighten properly; 3. Collision occurs during horizontal movement. 1) Replace PCB; 2) Replace motor; 3) Replace sensor; 4) Tighten properly; 4. Horizontal movement is not allowed in current vertical position: Home instrument or sample probe unit. Probe syringe movement error 1. Sensor status is incorrectCheck if belt or belt wheel is loose, tighten them if so; 2. Failed to find the zero position: 1) Remove obstacle or replace with a new assembly; 2) Replace PCB; 3) Replace motor; 4) Replace sensor; 5) Tighten properly; 10-4 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description incorrect. 2. Failed to find the zero position. Flag Probable Causes Corrective Actions A01007 Probe unit Warning Sample probe collides with an obstacle when aspiratingPosition:Sam ple ID/bar code:Position: / 1) Collision sensor or circuit or cable failure; 2) Collision sensor is not blocked when assembling the sample probe; 3) Incorrect sample probe positioning; 4) Incorrect sample/reagent carousel positioning. 1) Replace sensor or PCB or cable; 2) Reassemble; 3) Rotate the positioning sensor to remove dust or adjust the probe to be vertical; if problem remains, adjust the positioning parameters; 4) Adjust sample/reagent carousel positioning parameters. A01024 Probe unit Warning Insufficient sample Position: (diluting position on reaction carousel) Sample ID/bar code: / 1. No sample is loaded on the designated position; 2. Insufficient sample. 3. Sample probe level detection error. 1. Check if there is sufficient sample, and run the test again; 2. If there are level detection error, check sample probe, level detection board and connecting cable; replace if any abnormity is identified. (Probe level detection failed Position: (diluting position on reaction carousel) Sample ID/bar code:) A01028 Probe unit Error The probe detects no fluid level during washing. / 1. No deionized water or hydro error; 2. Sample probe level detection error. 1. Check if hydro is normal; troubleshoot if there is any abnormity; 2. If level detection error is identified, check probe, level detection board and connecting cable; replace if any abnormity is identified. A01033 Probe unit Warning The probe detects no fluid level on reaction carousel when dispensing.Cuvette No.:Sample ID/bar / 1. There is no reagent or insufficient reagent in the cuvette; 2. Incorrect vertical extreme 1. Check if R1 volume is sufficient and the reagent bottle is free of air bubbles, and then try again;2. Check if the vertical extreme position parameter for sample probe on reaction carousel is correct;3. If 10-5 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description code:Chemistry:(Prob e level detection failedCuvette No.:Sample ID/bar code:Chemistry:) Flag Probable Causes position parameter for sample probe on reaction carousel; 3. Sample probe level detection error. Corrective Actions there are level detection error, check sample probe, level detection board and connecting cable; replace if any abnormity is identified. A01039 Probe unit Error Instruction execute error. Specific error: Error details: / Instrument instructions cannot be executed. 1. Switch off the analyzer power and switch on it again. Recover failure by performing the Home maintenance procedure. 2. If this message appears for 3 times, contact our R&D department. A02007 Probe unit Warning Sample probe collides with an obstacle when aspirating Reagent position: Position: / 1) Collision sensor or circuit or cable failure; 2) Collision sensor is not blocked when assembling the sample probe; 3) Incorrect sample probe positioning; 4) Incorrect sample/reagent carousel positioning. 1) Replace sensor or PCB or cable; 2) Reassemble; 3) Rotate the positioning sensor to remove dust or adjust the probe to be vertical; if problem remains, adjust the positioning parameters; 4) Adjust sample/reagent carousel positioning parameters. A02023 Probe unit Warning Insufficient reagentChemistry:Posi tion:(Probe level detection failedChemistry:Positi on:) / 1. There is no reagent or insufficient reagent on the designated position; 2. Sample probe level detection error. 1. Check if the reagent is sufficient, and then try again.2. If there are level detection error, check sample probe, level detection board and connecting cable; replace if any abnormity is identified. A02025 Probe unit Error The probe dispenses insufficient reagent. Cuvette No.: Sample ID/bar code: Chemistry: / 1. The reagent probe aspirates nothing. 2. Syringe assembly error. 1. Check if the reagent meets the requirement and is sufficient in volume, and then try again. 2. Check if there is any abnormity of syringe assembly. A02027 Probe unit Warning Water residues exist in the cuvette / deionized water residual exists in cuvette. Check if aspirating tubing and solenoid valves of wash station is normal; BS-240 Chemistry Analyzer Service Manual 10-6 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Cuvette No.: Sample ID/bar code: Chemistry: Flag Probable Causes Corrective Actions Mixer vertical movement error 1. Sensor status error Belt or belt pulley is loose. 2. Failed to find the zero position. 1) Assembly is jammed; 2) Driving circuit error; 3) Vertical motor error; 4) Vertical sensor error; 5) Belt pulley is loose; 3. Vertical movement is not allowed in current horizontal position. Software error Mixer horizontal movement error 1. Sensor status error Belt or belt pulley is loose. 2. Failed to find the zero position. 1) Driving circuit error; 2) Motor error; 3) Home position sensor error; 4) Belt pulley is loose; 3. Horizontal movement is not Mixer vertical movement error 1. Sensor status error Check if belt or belt wheel is loose, tighten them if so; 2. Failed to find the zero position. 1) Remove obstacle or replace with a new assembly; 2) Replace PCB; 3) Replace motor; 4) Replace sensor; 5) Tighten properly; 3. Vertical movement is not allowed in current horizontal position. Reset the whole unit or the relevant unit. Mixer horizontal movement error 1. Sensor status error Check if belt or belt wheel is loose, tighten them if so; 2. Failed to find the zero position. 1) Replace PCB; 2) Replace motor; 3) Replace sensor; 4) Tighten properly; 3. Horizontal movement is not allowed in (Probe level detection failed Cuvette No.: Sample ID/bar code: Chemistry:) A05006 Mixer Error Mixer vertical movement error Position: Error: Mixer horizontal movement error Position: Error: Mixer vertical movement error 1. Sensor status is incorrect. 2. Failed to find the zero position. 3. Moving vertically is not allowed in current horizontal position. Mixer horizontal movement error 1. Sensor status is incorrect. 2. Failed to find the 10-7 / BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description zero position. 3. Moving horizontally is not allowed in current vertical position. Flag Probable Causes allowed in current vertical position. Software error Corrective Actions current vertical position. Reset the whole unit or the relevant unit. A05007 Mixer Unit Error Instruction execute error. Specific error: Error details: / Instrument instructions cannot be executed. 1. Switch off the analyzer power and switch on it again. Recover failure by performing the Home maintenance procedure. 2. If this message appears for 3 times, contact our R&D department. A06006 Reaction Carousel Unit Error Reaction carousel movement error Error: 1. Failed to find the zero position. 2. The coder missed steps 3. The reagent carousel inner ring missed steps when moving to the home position. / Reaction carousel movement error 1. Failed to find the zero position. Home position sensor error 2. The coder missed steps 1) Assembly is jammed; 2) Driving circuit error; 3) Motor error; 4) Sensor error; 5) Belt pulley is loose; 3. The reagent carousel inner ring missed steps when moving to the home position. Home position sensor error 1. Failed to find the home position; Replace sensor or cable 2. The coder missed steps 1) Remove obstacle or replace with a new assembly; 2) Replace PCB; 3) Replace motor; 4) Replace sensor or cable; 5) Replace belt pulley and motor assembly; 3. The reagent carousel inner ring missed steps when moving to the home position. Replace sensor or cable A06007 Reaction Carousel Unit Error Filter wheel movement error / 1. Software instruction execute error 2. The filter wheel motor goes wrong. 3. The home position sensor of the filter wheel goes wrong. 1. Switch off the analyzer power and switch on it again. Recover failure by performing the Home maintenance procedure. 2. Check if the cable of the filter wheel motor and the sensor are normal. 3. Check if the filter wheel motor and the BS-240 Chemistry Analyzer Service Manual 10-8 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions sensor are faulty. A06008 Reaction Carousel Unit Error Instruction execute error. Specific error: Error details: / Instrument instructions cannot be executed. 1. Switch off the analyzer power and switch on it again. Recover failure by performing the Home maintenance procedure. 2. If this message appears for 3 times, contact our R&D department. A07006 Sample/reag ent carousel unit Error Sample/Reagent carousel movement error. Error: 1. Failed to find the zero position. 2. The coder missed steps 3. The reagent carousel inner ring missed steps when moving to the home position. / Sample/Reagent carousel movement error. 1. Failed to find the zero position. Home position sensor error 2. The coder missed steps 1) Assembly is jammed; 2) Driving circuit error; 3) Motor error; 4) Sensor error; 5) Belt pulley is loose; 3. The reagent carousel inner ring missed steps when moving to the home position. Home position sensor error Sample/Reagent carousel movement error. 1. Failed to find the zero position. Replace sensor or cable 2. The coder missed steps 1) Remove obstacle or replace with a new assembly; 2) Replace PCB; 3) Replace motor; 4) Replace sensor or cable; 5) Replace belt pulley and motor assembly; 3. The reagent carousel inner ring missed steps when moving to the home position. Replace sensor or cable A07009 Sample/reag ent carousel unit Error Sample bar reader error / Sample bar code reader connection error due to system failure. If this occurs repeatedly, and failure recovery cannot be performed, bar code reader is not initialized correctly, or communication cable is disconnected. If bar code reader is not available, please try again. Recover the failure. If the error remains, initialize the sample bar code reader. If initialization cannot be performed, check if bar code reader is connected properly. If problem remains, replace bar code reader or control driver board. 10-9 code BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions A07010 Sample/reag ent carousel unit Warning Sample bar errorPosition code / Sample bar code reader does not work normally due to communication error. Perform the bar code scan operation again. If problem remains, check bar code reader connection, replace bar code reader and control driver board. A07011 Sample/reag ent carousel unit Error Sample bar code sending buffer is full / Sample bar code sending buffer is full due to communication error. Recover the failure analyzing unit. A09011 Sample/reag ent carousel unit Error Reagent bar code reader does not work normally / Reagent bar code reader connection error due to system failure. If this occurs repeatedly, and failure recovery cannot be performed, bar code reader is not initialized correctly, or communication cable is disconnected. If bar code reader is not available, please try again. Recover the failure. If the error remains, initialize the sample bar code reader. If initialization cannot be performed, check if bar code reader is connected properly. If problem remains, replace bar code reader or control driver board. A09012 Sample/reag ent carousel unit Warning Reagent bar code reader data error Reagent position: / Communication error. Reagent bar code instruction buffer is full. Perform the bar code scan operation again. If problem remains, check bar code reader connection, replace bar code reader and control driver board. A09014 Sample/reag ent carousel unit Error Reagent bar code sending buffer is full Reagent position: / Reagent bar code reader does not work normally due to communication error. Recover the failure analyzing unit. A11005 Wash unit Error Wash station movement error Error: 1. Movement error 2. The wash station collides with an obstacle when moving. / 1. Sensor status error Belt pulley is loose; 2. Home position is not found. 1) Driving circuit error; 2) Motor error; 3) Home position sensor error; 4) Belt pulley is loose; 3. The wash station collides with an obstacle when moving. 1. Sensor status error Check if belt or belt wheel is loose, tighten them if so; 2. Home position is not found. 1) Replace PCB; 2) Replace motor; 3) Replace sensor; 4) Tighten properly; 3. The wash station collides with an BS-240 Chemistry Analyzer Service Manual or or 10-10 reboot reboot the the 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes 1) Wash probes are not vertical; 2) Reaction carousel positioning is deviated; 3) Collision sensor or circuit error (Only BS-240) Corrective Actions obstacle when moving. 1) Align wash probes; 2) Align reaction carousel positioning; 3) Replace sensor, cable or PCB (Only BS-240) A11013 Wash unit Warning Water tank is empty / 1. Water tank is empty and not filled. 2. The low-level floater of the water tank goes wrong. 1. Check if the water level inside the water tank is low and replenish.2. Check if the error is accidental. 3. If the error is not accidental, contact our customer service department or your local distributor. A11015 Wash unit Warning Diluted wash solution tank is empty. / 1. Diluted wash solution tank is empty and not filled. 2. The low-level floater of the diluted wash solution tank goes wrong. (Only BS-240) 1. Replenish the diluted wash solution until the tank is full. 2. Check if the low-level floater signal of the diluted wash solution tank is correct. (Only BS-240) A11020 Wash unit Warning High concentration waste tank is full / 1. The high concentration waste tank is full 2. The floater of the high concentration waste tank goes wrong. (Only BS-240) 1. Check the high-concentration waste tank. If the tank is full, replace it. Cover the full tank and dispose of the waste properly. 2. Check if high-concentration waste goes wrong. (Only BS-240) A11034 Wash unit Error Deionized water syringe movement error. 1. Sensor status is incorrect. 2. Failed to find the mechanical zero position. / 1. Sensor status error. 2. The syringe assembly is probably forced to move. 3. Failed to find the zero position. 1) Assembly is jammed; 2) Driving circuit error; 3) Motor error; 4) Sensor error; 1. Replace sensor or cable; 2. Remove obstacle or replace with a new assembly; 3. Replace drive board. 4. Replace the motor; (Only BS-240) 10-11 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes (Only BS-240) Corrective Actions A11038 Wash unit Error Instruction execute error. Specific error: Error details: / Instrument instructions cannot be executed. 1. Switch off the analyzer power and switch on it again. Recover failure by performing the Home maintenance procedure. 2. If this message appears for 3 times, contact our R&D department. A12005 Temperature Control Unit Warning Reaction carousel temperature is out of range TDISP temperature: TS01: / 1. The ambient temperature is out of range. 2. The temperature sensor goes wrong. (component error, cable error) 3. The temperature protection switch goes wrong. (component error, cable error) 4. The heater goes wrong. (component error, cable error) 5. The fan goes wrong. (component error, cable error) 6. PCB error 7. Parameters are lost. 8. Electromagnetic interference exists. 1. Reproduce failure. 2. Detect ambient temperature to check if it is out of range. 3. Temperature sensor error: check for any signs of short circuit or broken circuit; check if connections are secure. 4. Temperature protection switch error: check if protection switch is shut and connections are secure. 5. Heater error: check if resistance under normal temperature is correct, and connections are secure. 6. Fan error: check if fan is working normally, and connections are secure. 7. PCB error 8. Parameters are lost: check if temperature control parameters are correct. 9. Electromagnetic interference exists. A12006 Temperature Control Unit Warning Temperature of wash solution for cleaning cuvettes is out of range Temperature: / 1. The ambient temperature is out of range. 2. The temperature sensor goes wrong. (component error, cable error) 3. The temperature protection 1. Reproduce failure. 2. Detect ambient temperature to check if it is out of range. 3. Temperature sensor error: check for any signs of short circuit or broken circuit; check if connections are secure. BS-240 Chemistry Analyzer Service Manual 10-12 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes switch goes wrong. (component error, cable error) 4. The heater goes wrong. (component error, cable error) 5. PCB error 6. Parameters are lost. 7. Electromagnetic interference exists. (Only BS-240) Corrective Actions 4. Temperature protection switch error: check if protection switch is shut and connections are secure. 5. Heater error: check if resistance under normal temperature is correct, and connections are secure. 6. PCB error 7. Parameters are lost: check if temperature control parameters are correct. 8. Electromagnetic interference exists. (Only BS-240) A12007 Temperature Control Unit Warning Temperature of deionized water for cleaning cuvettes is out of range Temperature: / 1. The ambient temperature is out of range. 2. The temperature sensor goes wrong. (component error, cable error) 3. The temperature protection switch goes wrong. (component error, cable error) 4. The heater goes wrong. (component error, cable error) 5. PCB error 6. Parameters are lost. 7. Electromagnetic interference exists. (Only BS-240) 1. Reproduce failure. 2. Detect ambient temperature to check if it is out of range. 3. Temperature sensor error: check for any signs of short circuit or broken circuit; check if connections are secure. 4. Temperature protection switch error: check if protection switch is shut and connections are secure. 5. Heater error: check if resistance under normal temperature is correct, and connections are secure. 6. PCB error 7. Parameters are lost: check if temperature control parameters are correct. 8. Electromagnetic interference exists. (Only BS-240) A12010 Temperature Control Unit Warning Reagent preheating temperature is out of range. / 1. The ambient temperature is out of range. 2. The temperature sensor goes wrong. (component error, cable error) 1. Reproduce failure. 2. Detect ambient temperature to check if it is out of range. 3. Temperature sensor error: check for any signs of short circuit or broken 10-13 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes 3. The temperature protection switch goes wrong. (component error, cable error) 4. The heater goes wrong. (component error, cable error) 5. PCB error 6. Parameters are lost. 7. Electromagnetic interference exists. Corrective Actions circuit; check if connections are secure. 4. Temperature protection switch error: check if protection switch is shut and connections are secure. 5. Heater error: check if resistance under normal temperature is correct, and connections are secure. 6. PCB error 7. Parameters are lost: check if temperature control parameters are correct. 8. Electromagnetic interference exists. A21001 Wash unit Error Wash solution syringe movement error. 1. Sensor status is incorrect. 2. Failed to find the mechanical zero position. / 1. Sensor status error. 2. The syringe assembly is probably forced to move. 3. Failed to find the zero position. 1) Assembly is jammed; 2) Driving circuit error; 3) Motor error; 4) Sensor error; (Only BS-240) 1. Replace sensor or cable; 2. Remove obstacle or replace with a new assembly; 3. Replace drive board. 4. Replace the motor; (Only BS-240) A22001 ISE module Error Slope out range.Electrode: / 1. Electrode is not installed correctly. 2. Calibration is invalid. 3. Electrode is degenerated. 4. Bubbles exist inside reference electrode. 5. Reference electrode has been used for a long time. 6. Electrodes interfere with each other. 7. Module or fluidic temperature 1. Remove and reinstall the electrode. 2. Replace the calibrator. 3. Replace the failed electrode and run tests again. 4. Remove the electrode, clap it to remove the bubbles, reinstall it, and then run calibration. 5. Replace the reference electrode and run tests again. 6. Replace electrode by different combinations to troubleshoot the failed BS-240 Chemistry Analyzer Service Manual of 10-14 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description A22002 ISE module Error Air in position: A22004 ISE module Error ISE unit cannot connected. A22005 ISE module A22006 ISE module 10-15 sample, Flag Probable Causes exceeds 32°C. Corrective Actions one. 7. Monitor the temperature. If the ambient temperature is too high, move the instrument to another position. 8. If the problem persists, contact Mindray. / 1. Sample is insufficient, or contains bubbles after being dispensed. 2. No or insufficient sample is dispensed to the injection port. 3. Leaks exist due to improper installation of electrode. 4. Waste pump tube is aging or broken. 1. Increase the sample volume. Place enough sample in the tube, at least 90μl for ISE measurement. 2. Electrode is not installed correctly. Reinstall it. 3. Check the waste tube, and if necessary, replace it. be / 1. Power supply of ISE module goes wrong. 2. Communication cable between ISE module and middle-layer unit goes wrong. 3. Communication interface or pins go wrong. 4. Main control board of the ISE module goes wrong. 1. Replace the 24V power supply board of the ISE module. 2. Replace the communication cable. 3. Replace the interface or pins. 4. Replace the main control board of the ISE module. Error ISE unit response error / 1. Communication cable between ISE module and middle-layer unit goes wrong. 2. Communication interface or pins go wrong. 3. Main control board of the ISE module goes wrong. 4. Software error 1. Replace the communication cable. 2. Replace the interface or pins. 3. Replace the main control board of the ISE module. 4. Upgrade or reinstall the software. Error Purge A and B failed. / 1. Leaks exist due to improper installation of electrode or O 1. Reinstall the electrode and check for the O ring. BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes ring's missing. 2. Sample injection port of electrode inside is blocked. 3. Calibrator in the reagent pack is exhausted. 4. Purge cycle is not enough. 5. Peristaltic pump tube is aging, blocked or broken. 6. Calibrator cannot be output due to the blocked wand tube. Corrective Actions 2. Use warm water to rinse the sample injection port, and unclog the electrode tube. Check the inside of the reference electrode for salt buildup. 3. Replace the reagent pack. 4. Increase the prime cycle. 5. Replace the pump tube. 6. Use warm water to unclog the wand tube. A22007 ISE module Warning ISE reagent is going to be exhausted. / Calibrator is exhausted. Replace the reagent pack with a new one. A22008 ISE module Error Electrode Overflow Electrode: Voltage / 1. Electrode or reagent pack is invalid. 2. Electrode is unsteady. 3. New reagent pack is unstable. 4. Reference electrode has been used for over 6 months. 5. Main control board of the ISE module goes wrong. 1. Replace the failed electrode or reagent pack. 2. New electrode can become stable only after 15 minutes since installation. 3. Multiple calibrations are required after installing new reagent pack. 4. Replace the reference electrode. 5. Replace the main control board. A22009 ISE module Error Electrode slope drift. Electrode voltage drift.Electrode: / 1. Electrode or reagent pack is invalid. 2. Electrode is unsteady. 3. New reagent pack is unstable. 4. Reference electrode has been used for over 6 months. 5. Main control board of the ISE module goes wrong. 6. The ambient temperature is fluctuating greatly. 1. Replace the failed electrode or reagent pack. 2. New electrode can become stable only after 15 minutes since installation. 3. Multiple calibrations are required after installing new reagent pack. 4. Replace the reference electrode. 5. Replace the main control board. 6. Monitor the ambient temperature and ensure the fluctuation is within ±4. BS-240 Chemistry Analyzer Service Manual 10-16 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Probable Causes Corrective Actions / 1. Electrode failure. 2. Environment interference. 3. Main control board of the ISE module goes wrong. 4. Fluidic leaks exist, resulting in salt buildup around the electrode or tube. 1. Replace the electrode. 2. Relocate the instrument. 3. Replace the main control board of the ISE module. 4. Clean the tube and electrode. A22010 ISE module Error Electrode Noise Electrode: A22011 ISE module Error Air in calibrator B / 1. Calibrator B is exhausted. 2. Calibrator B tube contains many air bubbles. 3. Peristaltic pump B tube is aging, clogged and broken. 4. Waste pump tube is aging, clogged and broken. 5. The sample injection port and fluidic tube is blocked or leaking. 6. The air bubble detector is invalid. 1. Replace the reagent pack with a new one. 2. Perform purge B to remove air bubbles. 3 and 4. Replace the peristaltic pump tube. 5. Clean the sample injection port, or reinstall the electrode. 6. Replace the air bubble detector. A22012 ISE module Error Air in calibrator A / 1. Calibrator A is exhausted. 2. Calibrator A tube contains many air bubbles. 3. Peristaltic pump A tube is aging, clogged and broken. 4. Waste pump tube is aging, clogged and broken. 5. The sample injection port and fluidic tube is blocked or leaking. 6. The air bubble detector is invalid. 1. Replace the reagent pack with a new one. 2. Perform purge A to remove air bubbles. 3 and 4. Replace the peristaltic pump tube. 5. Clean the sample injection port, or reinstall the electrode. 6. Replace the air bubble detector. A22013 ISE module Error Pump calibration result error / 1. Peristaltic pump tube is aging. 2. Aspiration and dispensing of sample probe goes wrong. 1. Replace the pump tube. 2. Replace the sample probe. 10-17 Voltage Flag BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description A22014 ISE module Error Bubble failure. A22015 ISE module Error Reagent pack reading error. the reagent again. A22016 ISE module A22017 Probable Causes Corrective Actions / 1. Air bubble detector board is eroded due to the leaks at the joint of sample injection port and bubble detector. 2. The air bubble detector is invalid. Replace the air bubble detector. chip Load pack / 1. Reagent pack is not installed. 2. Reagent pack wand is failed. 1. Install the reagent pack. 2. Install the reagent pack wand. Error Cannot write into the chip of reagent pack. / 1. Reagent pack is not installed. 2. Reagent pack wand is failed. 1. Install the reagent pack. 2. Install the reagent pack wand. ISE module Error Air in solution wash / 1. ISE wash solution is insufficient. 2. Waste pump tube is aging, clogged and broken. 3. The sample injection port and fluidic tube is blocked or leaking. 4. The air bubble detector is invalid. 1. Place sufficient ISE wash solution. 2. Replace the peristaltic pump tube. 3. Clean the sample injection port, or reinstall the electrode. 4. Replace the air bubble detector. A22018 ISE module Error The ISE tube has no liquid or has leaks. / 1. ISE wash solution is insufficient. 2. Waste pump tube is aging, clogged and broken. 3. The sample injection port and fluidic tube is blocked or leaking. 4. The air bubble detector is invalid. 1. Place sufficient ISE wash solution. 2. Replace the peristaltic pump tube. 3. Clean the sample injection port, or reinstall the electrode. 4. Replace the air bubble detector. A22019 ISE module Error Stored calibration value error. / 1. Communication cable is disconnected. 2. The interface and pins are faulty. 3. Main control board of the 1. Replace the communication cable. 2. Replace the interface or pins. 3. Replace the main control board of the ISE module. BS-240 Chemistry Analyzer Service Manual detector Flag ISE 10-18 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes module goes wrong. 4. Unmatched software version. Corrective Actions 4. Upgrade or reinstall the software. A22021 ISE module Error Command format or execution error / 1. Communication cable is disconnected. 2. The interface and pins are faulty. 3. Main control board of the module goes wrong. 4. Unmatched software version. 1. Replace the communication cable. 2. Replace the interface or pins. 3. Replace the main control board of the ISE module. 4. Upgrade or reinstall the software. A22022 ISE module Error Air in segment / 1. ISE wash solution is insufficient. 2. Waste pump tube is aging, clogged and broken. 3. The sample injection port and fluidic tube is blocked or leaking. 4. The air bubble detector is invalid. 1. Place sufficient ISE wash solution. 2. Replace the peristaltic pump tube. 3. Clean the sample injection port, or reinstall the electrode. 4. Replace the air bubble detector. A22023 ISE module Error No reagent module has been loaded. / 1. Reagent pack is not installed. 2. Reader/writer error. 1. Install the reagent pack. 2. Install the reagent pack wand. A22024 ISE module Error ISE response code error / 1. Communication cable is disconnected. 2. The interface and pins are faulty. 3. Main control board of the module goes wrong. 4. Unmatched software version. 1. Replace the communication cable. 2. Replace the interface or pins. 3. Replace the main control board of the ISE module. 4. Upgrade or reinstall the software. A22027 / Error Fluidic prime failed. / Fluidic prime is not performed. Switch off the analyzing unit power and switch on it again. If this message appears for 3 times, contact our customer service department or your local distributor. 10-19 check BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions A22036 / Error Initializing sample bar code reader failed. / Sample bar code reader connection error due to system failure. If this occurs repeatedly, and failure recovery cannot be performed, bar code reader is not initialized correctly, or communication cable is disconnected. Recover the failure. If the error remains, initialize the sample bar code reader. If initialization cannot be performed, check if bar code reader is connected properly. If problem remains, replace bar code reader or control driver board. A22038 / Error Scanning reagent bar code failed. / Scanning reagent bar code failed. Contact our R&D department. A22039 / Error Unmatched version. / 1. Executing version query instruction failed. 2. Returned control software version and that stored in the operating software does not match. 1. Switch off the analyzing unit power and switch on it again. 2. If this message appears for 3 times, contact our customer service department or your local distributor. C00007 Operating system Error CPU error / The CPU is too busy. Reboot the computer and operating software. If this message appears for 3 times, contact our customer service department or your local distributor. C00011 Operating system Error The operating software has not been closed normally. / The operating software abnormal or the analyzer powered off accidentally. is is After rebooting the software, perform the special wash procedure. C00012 Operating system Warning Sound card failure / No sound card is installed. Sound card failure. Incorrect sound card driver. Reinstall the sound card or the sound card driver. C00013 Operating system Error The last abnormal exit may cause absence of update of the reaction cuvette status. Check the reaction cuvette manually and replace it with a new cuvette to ensure accurate / The operating software did not exit normally for reasons such power failure. (Only BS-230) Remove and check the idle reaction cuvette segment on the reaction carousel. If the cuvette segment has been used, replace it; or, if it is not sure whether the cuvette segment has been used, still replace it. (Only BS-230) BS-240 Chemistry Analyzer Service Manual software 10-20 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description results. Flag Probable Causes Corrective Actions C01001 System communicati on Error Equipment cannot be connected / The serial cable connected; or the power is switched off. is not analyzer Check the serial port connection. Replug the cable. Check if the analyzing unit are powered on. Start the initialization again. Restart the computer and analyzing unit. If three continuous attempts are failed, contact our customer service department or your local distributor. C02001 Database Error Database failed initializing / Database software is not installed; database is not established; database system or database file is damaged or lost; database is opened exclusively Install database software; re-establish database; if problem remains, contact our R&D department C02002 Database Error Database failed upgrade / Database upgrade failed Install database software; re-establish database; if problem remains, contact our R&D department C02004 Database Warning Data backup failed / Database record is being used. Database is locked. Back up database again; if problem remains, contact our R&D department C02005 Database Warning Reading/Writing database failed / Cannot read or write necessary information from or into database Check if database has been damaged; contact our R&D department C03001 Result calculation Warning Result cannot be calculatedSample ID/bar code:Position:Chemist ry: RCE Absorbance data for calculation is incomplete, or the dividend is 0. Investigate the cause of photoelectric data loss; contact our R&D department C03002 Result calculation Warning Absorbance out of range Sample ID/bar code: Position: ABS The absorbance measured at the primary and secondary wavelength is greater than 3.3A. 1. Check the reaction curve. If R1 absorbency is too high, check whether the reagent has bubbles or the syringe leaks. 2. Check if the reagent is installed correctly. 10-21 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions 3. Rerun the test after diluting the reagent. 4. Contact our customer service department or your local distributor. C03003 Result calculation Warning R1 blank absorbance out of range Sample ID/bar code: Position: Chemistry: RBK The reagent goes wrong; the cuvette is not clear; the reaction cuvette is overflowed; or insufficient reagent is dispensed.(BS-240) The reagent goes wrong; the reagent is not replenished enough.(BS-230) Check if cuvette is clean or overflowed; check if reagent is sufficient, without air bubbles; check if chemistry parameters are reasonable; if previous problems do not exist, replace reagent and rerun the test (BS-240). Check if reagent is sufficient (if reagent bottle contains air bubbles); check if chemistry parameters are reasonable; if previous problems do not exist, replace reagent and rerun the test (BS-230). C03004 Result calculation Warning Substrate depletion Sample ID/bar code: Position: Chemistry: BOE The sample concentration is too high, and substrate depletion occurs during fixed-time measurements. Check the reaction curve and the substrate depletion limit. Rerun the test with diluted sample. C03005 Result calculation Warning Result cannot be calculatedSample ID/bar code:Position:Chemist ry: ENC The sample concentration is too high, and substrate depletion occurs within the lag time of rate check measurements. Check the reaction curve and the substrate depletion limit. Rerun the test with diluted sample. C03006 Result calculation Warning Linearity limit out of range Sample ID/bar code: Position: Chemistry: LIN The measuring points for result calculation are nonlinear, because the sample concentration is too high, or the substrate depletion limit is not specified or unreasonable. Check the reaction curve and the substrate depletion limit. Rerun the test with diluted sample. C03007 Result calculation Warning Prozone check error Sample ID/bar code: Position: PRO Antigen excess occurs due to too high sample concentration. Check the reaction curve and the prozone check parameters. Rerun the test with diluted sample. BS-240 Chemistry Analyzer Service Manual 10-22 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Chemistry: Flag Probable Causes Corrective Actions C03008 Result calculation Warning Sample concentration is higher than that of the highest-level calibrator Sample ID/bar code: Position: Chemistry: RRN The sample concentration exceeds the high limit of the calibrator concentration. Rerun the test with diluted sample. C03009 Result calculation Warning Mixed blank absorbance out of rangeChemistry: MBK The reagent goes wrong; the cuvette is not clear; the reaction cuvette is overflowed; or insufficient reagent is dispensed. Check if the cuvette is clear and not overflowed, the reagent is sufficient without air bubbles, and the chemistry parameters are reasonable. If yes, replace the reagent and then rerun the test. C03010 Result calculation Warning Blank response out of range Chemistry BLK The reagent goes wrong; insufficient reagent is dispensed; the cuvette contains air bubbles; the light drifts; or the cuvette is overflowed.(BS-240) The reagent goes wrong; insufficient reagent is dispensed; the cuvette contains air bubbles; the light drifts.(BS-230) Check if the cuvette is not overflowed, the reagent is sufficient without air bubbles, the light does not drift and the chemistry parameters are reasonable. If yes, replace the reagent and then rerun the test (BS-240). Check if reagent is sufficient (if reagent bottle contains air bubbles); check if chemistry parameters are reasonable; if the light drifts, if previous problems do not exist, replace reagent and rerun the test (BS-230). C03011 Result calculation Warning Calibration repeatability range. Chemistry: DUP The difference between the maximum and minimum response of the calibrator exceeds the specified limit. Check if the acceptance limit is reasonable, troubleshoot the error, and then recalibrate. Result calculation Warning SEN The difference of final response of the maximum and minimum concentration calibrators Check if the acceptance limit is reasonable and the reagent and calibrator are normal, and then recalibrate. C03012 10-23 out of Calibration sensitivity out of range. Chemistry: BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes exceeds the specified limit. Corrective Actions C03013 Result calculation Warning Calibration curve standard deviation out of range. Chemistry: CSD The calculated standard deviation of the calibration curve exceeds the specified limit. Check if the acceptance limit is reasonable and the reagent and calibrator are normal, and then recalibrate. C03014 Result calculation Warning Calibration determination coefficient out range Chemistry: DET The calculated fit of the calibration curve exceeds the specified limit. Check if the acceptance limit is reasonable and the reagent and calibrator are normal, and then recalibrate. of C03015 Result calculation Warning Calibration coefficient differential limit exceeds limit. Chemistry: FAC The slope difference is applicable to linear calibration only and refers to the K factor (slope) difference between two adjacent calibrations. It exceeds the specified limit. Check if the acceptance limit is reasonable and the reagent and calibrator are normal, and then recalibrate. C03016 Result calculation Warning Calibration curve not monotonic. Chemistry: MON The calibration data calibration curve are monotonic. and not Check if the calibrator is defined and placed correctly, and then recalibrate. C03017 Result calculation Warning Calibration curve is not convergent.Chemistry: COV For nonlinear calibration, a satisfying base cannot be calculated and no calibration curve is drawn. Check that the reagent and calibrator are normal, and then recalibrate. If troubleshooting fails, choose other calibration rules and recalibrate. C03018 Result calculation Warning Chemistry: Control: 1-2S warning 12s The QC result is between ±2 and ±3 standard deviations from the assigned mean concentration. No actions are required. C03019 Result calculation Warning Chemistry: Control: 1-3S out of control 13s The current QC result is greater than ±3 standard deviations from the assigned mean concentration. Check if the reagent is qualified, control sample is normal, and the instrument is working correctly. BS-240 Chemistry Analyzer Service Manual 10-24 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions C03020 Result calculation Warning Chemistry: Control: 2-2S out of control 22s Results of two controls in the same run or two continuous results of a control are on the same side and greater than ±2 standard deviations from the assigned mean concentration. Check if the reagent is qualified, control sample is normal, and the instrument is working correctly. C03021 Result calculation Warning Chemistry: Control: R-4S out of control R4s One result of a run is greater than +2 standard deviations from the assigned mean and the other greater than -2SDs. Check if the reagent is qualified, control sample is normal, and the instrument is working correctly. C03022 Result calculation Warning Chemistry: Control: 4-1S out of control 41s Results of two runs (4 results), or 4 continuous results of a control are on the same side and greater than ±1 standard deviation from the assigned mean concentration. Check if the reagent is qualified, control sample is normal, and the instrument is working correctly. C03023 Result calculation Warning Chemistry: Control: 10-X out of control 10x Results of recent 5 runs, or 10 continuous results of a control are on the same side. Check if the reagent is qualified, control sample is normal, and the instrument is working correctly. C03024 Result calculation Error Test period timed out; Cannot continue / 1. Software error 2. Operating system error Rerun the test, or reboot the operating software and analyzing unit. If the error remains, contact our customer service department or your local distributor. C03026 Result calculation Warning Photoelectric data is lost / Communication error. If the error persists, contact our R&D department. C03030 Result calculation Error Photoelectric measurement is time out Sample ID/bar code: Position: TD Software error 1. Rerun the operating software. 2. Restart the analyzing unit. 3. If problem remains, contact our R&D department. 10-25 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Chemistry: Flag Probable Causes Corrective Actions C03031 Result calculation Error Multiple photoelectric measurements are time outSample ID/bar code: Position: / Software error 1. Rerun the operating software. 2. Restart the analyzing unit. 3. If problem remains, contact our R&D department. C04001 Sample code bar Warning Duplicate sample bar code. Sample ID/bar code: %s Position 1: %s Position 2: %s / Duplicate bar code is used. Replace the duplicate sample bar code label. C04002 Sample code bar Warning Corresponding program information does not exist Sample ID/bar code: Position (sample/reagent carousel position) / The sample of the bar code has not been programmed. Program the sample of the bar code. C04006 Sample code bar Warning Sample is expired. Sample ID/bar code: Position (sample/reagent carousel position) / The sample is loaded after its shelf life is exceeded. The sample is expired. Replace the sample and program it again. Reject the expired sample. If the sample shelf life is too short, change it to a reasonable one. C04008 Sample code bar Warning Sample bar code is longer than 27 digits. Position (sample/reagent carousel position) / The bar code length is greater than the maximum value of 27 digits. Redefine the bar code with no more than 27 digits. C04009 Sample code bar Warning Sample bar code is less than 3 digits. Sample bar code: , Position: / The bar code length is less than the minimum value of 3 digits. Redefine the bar code with no less than 3 digits. C04012 Sample bar Warning Sample / Bar code data does not accord Reset the bar code format, or reprint and BS-240 Chemistry Analyzer Service Manual bar code 10-26 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component code Event Class Problem Description analysis error Sample bar code: Position: Flag Probable Causes with the defined format. Corrective Actions scan the bar code. C05001 Reagent bar code Warning Duplicate reagent bar code Position 1: Position 2: / Incorrect reagent or reagent bar code is being used, or an invalid reagent bar code is being used in a closed-reagent system. Closedreagent bar code is aligned with closed reagents, and cannot be used again for new reagent when a reagent is exhausted. Reprint the reagent bar code, or replace the reagent bottle with an invalid bar code. C05002 Reagent bar code Warning Reagent bar code information error. Position: / Incorrect reagent bar code is being used, or reagent bar code is not configured reasonably. The reagent bar code contains incomplete or incorrect reagent information, such as expiration date, reagent volume, etc. Print the new reagent bar code with correct settings and check the bar code against the settings. In the case of a closed-reagent system, replace the reagent bottle, or contact the reagent supplier. C05003 Reagent bar code Warning Reagent bar code analysis error.Position: / Incorrect reagent bar code is being used, or reagent bar code settings are incorrect. Nonclosed reagent bar code is being used in a closed-reagent system. The system fails to extract reagent information from the bar code. Check the reagent bar code settings, or reprint the reagent bar code against the settings. In the case of a closed-reagent system, contact the reagent supplier. C05006 Reagent bar code Error Wash solution position on sample/reagent carousel is occupied by another reagent. Position: / Reagent rather than wash solution is placed in the fixed wash solution position on sample/reagent carousel. Reposition the reagent, or remove it from the fixed wash solution position. C05008 Reagent bar code Error Physiological saline position on sample/reagent carousel is occupied by another reagent. Position: / Reagent rather than physiological saline is placed in the fixed physiological saline position on sample/reagent carousel. Reposition the reagent, or remove it from the fixed wash solution position. 10-27 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions C05010 / Warning / / The pretreat reagent position is occupied by other reagent. (Only BS-240) Check if the reagent is placed correctly and if the reagent position is set correctly (Only BS-240). C05011 / Warning / / The ISE wash solution position is occupied by other reagent. Check if the reagent is placed correctly and if the reagent position is set correctly. C06001 LIS communicati on Error LIS initialization error / Host file is damaged or does not exist. Reboot operating software; initialize LIS module C06002 LIS communicati on Error LIS communication parameter error / Host parameters error Check system configuration parameters; re-connect LIS C06003 LIS communicati on Error LIS error communication / Communication failure If the error occurs accidentally, send or receive the instruction again. If the error occurs for several times, contact LIS vendor. If problem remain, contact our R&D department. C06004 LIS communicati on Error LIS host cannot be connected / Abnormal network connection, or the LIS host is not started. Check if LIS host and LIS station is started normally. C06005 LIS communicati on Warning Sending sample results failed. Sample ID/bar code: Position: / Communication failure If the error occurs accidentally, send or receive the instruction again. If the error occurs for several times, contact LIS vendor. If problem remain, contact our R&D department. C06006 LIS communicati on Error Sending sample information failed. Sample ID/bar code: Position: / Communication failure If the error occurs accidentally, send or receive the instruction again. If the error occurs for several times, contact LIS vendor. If problem remain, contact our R&D department. BS-240 Chemistry Analyzer Service Manual 10-28 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions C06007 LIS communicati on Warning Inquiring sample information failed Sample ID/bar code: Position: / LIS server failure. If the error occurs accidentally, neglect it. Neglect the error. If the error occurs for several times, contact LIS vendor. If the problem remains, contact our customer service department or your local distributor. C06008 LIS communicati on Warning Downloading sample failed. Sample ID/bar code: Position: / Incorrect channel settings; or insufficient or redundant chemistries on the LIS host. Check and re-set the chemistry correspondence between the operating software and the LIS host. C07003 Light Source Error Light intensity is too weak / 1. The lamp is not installed correctly. 2. The cuvette is contaminated. 3. The lamp is aging. 4. The wash station dispenses liquid incorrectly (only BS-240). 1. Check if the lamp is installed correctly. 2. Perform the special wash procedure and then the lamp check procedure (only BS-240). 3. Replace the lamp. 4. Check if the wash station dispenses liquid with correct volume to reaction cuvettes (only BS-240). 5. Check if the cables on the AD collection board are installed correctly. 6. Replace the AD collection board. 7. If the problem remains, contact the R&D department of Mindray. C07004 Light Source Warning Cuvette blank out of range Cuvette No.: XX / 1. The cuvette is contaminated. 2. The lamp is aging. 3. The lamp is not installed correctly. 4. The wash station dispenses liquid incorrectly. 5. The photoelectric collection board goes wrong. (Only BS-240) 1. Replace or clean the failed cuvette. 2. Check if the lamp is installed correctly. 3. Perform the diluted wash procedure and then the cuvette check procedure. 5. Replace the lamp. 6. Check if the wash station dispenses liquid with correct volume to reaction cuvettes. 6. Check if the cables on the AD collection board are installed correctly. 7. Replace the AD collection board. 10-29 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions 8. If the problem remains, contact the R&D department of Mindray. (Only BS-240) C07005 Light Source Error Lamp is not turned on / 1. The lamp is damaged. 2. The lamp cable is not connected properly. 3. The power board of the lamp is not connected properly. 4. The power supply of the analyzing unit is disconnected. 5. The photoelectric collection board goes wrong. 1. Open the reaction carousel and check if the lamp is turned on. If it is not, rerun the operating software.2. Check if the lamp cable is tightened.3. Replace the lamp.4. Check if the connect of the lamp power board is loose, and if necessary, reinsert the connector.5. Check if the cables on the AD collection board are installed correctly.6. Replace the AD collection board.7. If the problem remains, contact the R&D department of Mindray. C07006 Light Source Error Light intensity is too strong / 1. A cuvette position has no cuvette installed. 2. The circuit gain is too high and beyond the measurement range. 1. Check if all cuvette positions have cuvettes installed. 2. Readjust the gain or(/and) lamp voltage. 3. If the problem remains, contact the R&D department of Mindray. C07007 Light Source Error Dark current is too high / 1. The circuit gain is too high and beyond the measurement range. 2. The power board of the lamp is not connected properly. 3. The photoelectric collection board goes wrong. 1. Perform dark current fluctuation test to identify failure. 2. Check if the abnormal channel has something to do with the channel gain. If it does, adjust the channel gain. 3. Check if the cables on the AD collection board are installed correctly. 4. Unplug the cables from the AD collection board and pre-amplification board and perform the dark current fluctuation test again. Check if the failure is related to the pre-amplification board. If yes, replace the optical BS-240 Chemistry Analyzer Service Manual 10-30 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions measurement assembly. 5. Replace the AD collection board. 6. If the problem remains, contact the R&D department of Mindray. C07009 Light Source Warning Water blank out of range (10X). L! 1. Overflow (only BS-240). 2. The lamp has been replaced incorrectly. 3. Cuvette Check is not performed after cuvette maintenance. 4. Wash solution inside the cuvette is little (only BS-240). 5. The cable connector or retaining screw of the lamp are not tightened. 6. The lamp is aging. 7. The photometer goes wrong. 1. Check if the cuvette is overflowing (only BS-240). 2. Check if the Replace Lamp command is executed during lamp replacement. 3. Check if the Cuvette Check command is executed after cuvette maintenance. 4. Check if the cleaning liquid inside the cuvette is no less than half of the cuvette (only BS-240). 5. Check if the cable connectors and retaining screw of the lamp have been tightened. 6. Check if the reaction curve fluctuates irregularly. If yes, replace the lamp. 7. If the problem remains, contact the R&D department of Mindray. C07012 Others Warning Storage device error; Cannot import data / No floppy disk or U disk is inserted. No file is found in the floppy disk or U disk, or file error, or file is damaged. The floppy disk or U disk is locked or damaged. Check if a USB drive or floppy disk is inserted or full. Check if the storage device is damaged. 10-31 BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions C07013 Others Warning Storage device error. Cannot export data / No floppy disk or U disk is inserted. Insufficient disk space. The floppy disk or U disk is locked or damaged. Check if a USB drive or floppy disk is inserted or full. Check if the storage device is damaged. C07014 Others Warning Biochemistry exhausted Chemistry: Position: reagent / All reagents of the reagent type for the chemistry are less than the minimum limit. All reagents of the type are too little to be detected. Refill or replace the reagent. C07016 Others Warning Insufficient wash solution. Position: / Insufficient wash solution on the sample/reagent carousel. Refill the wash solution sample/reagent carousel. on the C07017 Others Warning Wash solution exhausted. Position: is / The wash solution on the sample/reagent carousel is exhausted. Refill the wash solution sample/reagent carousel. on the C07022 Others Warning Less than X tests are left in biochemistry reagent Chemistry: / All reagents of the reagent type for the chemistry are less than the minimum limit. All reagents of the type are too little to be detected. Refill or replace the reagent. C07023 Others Warning Chemistry:%s, 30 minutes left for next calibration. / The calibration factors are about to be expired. Recalibrate the chemistry. BS-240 Chemistry Analyzer Service Manual 10-32 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes Corrective Actions C07027 Others Warning Calibrator %s has been expired / The calibrator is expired. Replace the calibrator. C07028 Others Warning Chemistry: %s, lot No.: %s,position: %s, has been expired / Expired reagent Replace the reagent. C07029 Others Warning Chemistry: %s, No.: %s, position: %s, exceeded uncapping time / The uncapping time of reagent pack is too long. lot the Replace the reagent. has the C07034 Others Warning Insufficient physiological saline on sample/reagent carousel Position: / Insufficient physiological saline on sample/reagent carousel Refill the physiological saline on the sample/reagent carousel. C07035 Others Warning Physiological saline on sample/reagent carousel is exhausted. Position: / Physiological sample/reagent exhausted Refill the physiological saline on the sample/reagent carousel. C07036 Others Warning Chemistry:%s. Calibration factors are expired / The calibration factors have been expired. Recalibrate the chemistry. C07037 Others Warning Reagent bottle number of %s chemistry is changed. Please recalibrate / Serial number of the reagent is changed. Reagent serial number of the chemistry is changed. Please recalibrate C07038 Others Warning Reagent bottle number of %s chemistry is changed. Please recalibrate / Lot number of the reagent is changed. Reagent lot number of the chemistry is changed. Please recalibrate 10-33 saline carousel on is BS-240 Chemistry Analyzer Service Manual 10 Troubleshooting 10.1 Error Codes and Troubleshooting Error Code Component Event Class Problem Description Flag Probable Causes C07039 Others Warning Calibration factors of %s chemistry are expired. Recalibrate / The calibration expired. C07040 Others Warning Biochemistry exhausted. Chemistry: / Reagent is exhausted. Reagent level cannot detected. reagent factors Corrective Actions are The calibration factors are expired. Refill or replace the reagent. be C07041 Others Error Less than X tests are left in ISE reagent / Inventory of the ISE reagent is lower than the alarm limit. Check the inventory, reagent if necessary. C07042 Others Warning %s, Lot number:%s, position: %s, has been expired / One or more special reagents have expired. Replace the reagent. C07043 / Warning / / Pretreat reagent is exhausted (only BS-240) Replenish pretreat reagent (only BS-240) C07044 / Warning / / Pretreat reagent (only BS-240) Replenish pretreat reagent (only BS-240) c / Warning / / ISE wash solution is exhausted. Replenish ISE wash solution C07046 / Warning / / Insufficient ISE wash solution. Replenish ISE wash solution BS-240 Chemistry Analyzer Service Manual is deficient and 10-34 refill ISE 10 Troubleshooting 10.2 Error feedback sheet 10.2 Error feedback sheet BS-240 Analyzer Error Feedback Sheet Serial No. Hospital Name Installation date Error date Software version Customer service engineer and Tel Problem Description Error content Error process (for software problems, describe detailed operation steps, preferably give screenshot). Error occurrence time (yy/mm/dd/hh/mm; give software alarm reporting time if any) Error description: Software alarm code: Error frequency Always; repeatable; frequent: times per hour; other On-site Corrective Actions BS-240 Chemistry Analyzer Service Manual occasional: times per day; not Effects 10-35 10 Troubleshooting 10.2 Error feedback sheet Document The customer service engineer should feed back the following documents: AlarmFile corresponding to the error date; Address: D:\Mindray\BS240\OperationSoft\AlarmFile; LOG file corresponding to the error date; Address: D:\Mindray\BS240\OperationSoft\WorkstationLOG; D:\Mindray\BS240\OperationSoft\Instrument_1\InstrumentLOG; 3) Provide screenshots in the case of software issues. 4) If abnormal halt or database error occurs, return the database file in the following path: Address: D:\Mindray\BS240\OperationSoft\DataBase\Backup (Copy the file under this folder after restarting the computer) 5) If a failure leading to abnormal clinical results occurs, return the database file in the following path as well as the reaction curve data of the failure date: Address: D:\Mindray\BS240\OperationSoft\DataBase\Backup (Copy the file under this folder after restarting the computer) D:\Mindray\BS240\OperationSoft\AbsData Note: Do not export the logs through the software screen. R&D response The R&D engineer confirms the foregoing faults and gives corrective measures and plans. BS-240 Chemistry Analyzer Service Manual P/N:046-009169-00(1.0)