ASME BPVC.V-2021 SECTION V 2021 ASME Boiler and Pressure Vessel Code An International Code N on de st ruct iv e Exa min a t ion ASME BPVC.V-2021 ARTICLE 1 SUBSECTION A NONDESTRUCTIVE METHODS OF EXAMINATION ARTICLE 1 GENERAL REQUIREMENTS T-110 SCOPE T-120 (a) This Section of the Code contains requirements and methods for nondestructive examination (NDE), which are Code requirements to the extent they are specifically referenced and required by other Code Sections or referencing documents. These NDE methods are intended to detect surface and internal imperfections in materials, welds, fabricated parts, and components. They include radiographic examination, ultrasonic examination, liquid penetrant examination, magnetic particle examination, eddy current examination, visual examination, leak testing, and acoustic emission examination. See Nonmandatory Appendix A of this Article for a listing of common imperfections and damage mechanisms, and the NDE methods that are generally capable of detecting them. (b) For general terms such as inspection, flaw, discontinuity, evaluation, etc., refer to Mandatory Appendix I. (c) New editions of Section V may be used beginning with the date of issuance and become mandatory 6 months after the date of issuance unless modified by the referencing document. (d) Code Cases are permissible and may be used, beginning with the date of approval by ASME. Only Code Cases that are specifically identified as being applicable to this Section may be used. At the time a Code Case is applied, only the latest revision may be used. Code Cases that have been incorporated into this Section or have been annulled shall not be used, unless permitted by the referencing Code. Qualifications using the provisions of a Code Case remain valid after the Code Case is annulled. The Code Case number shall be listed on the NDE Procedure or Personnel Certification, as applicable. GENERAL (a) Subsection A describes the methods of nondestructive examination to be used if referenced by other Code Sections or referencing documents. (b) Subsection B lists Standards covering nondestructive examination methods which have been accepted as standards. These standards are not mandatory unless specifically referenced in whole or in part in Subsection A or as indicated in other Code Sections or referencing documents. Where there is a conflict between Subsection A and Subsection B, the requirements of Subsection A take precedence. (c) Any reference to a paragraph of any Article in Subsection A of this Section includes all of the applicable rules in the paragraph. In every case, reference to a paragraph includes all the subparagraphs and subdivisions under that paragraph. NOTE: For example, a reference to T-270 includes all of the rules contained in T-271 through T-277.3. (d) Reference to a standard contained in Subsection B is mandatory only to the extent specified. NOTE: For example, T-233 requires that Image Quality Indicators be manufactured and identified in accordance with the requirements or alternatives allowed in SE-747 or SE-1025, and Appendices, as appropriate for the style of IQI to be used. These are the only parts of either SE-747 or SE-1025 that are mandatory in Article 2. In many cases, Subsection B documents are not mandatory and are intended only for guidance or reference use. (e) For those documents that directly reference this Article for the qualification of NDE personnel, the qualification shall be in accordance with their employer’s written practice which shall be in accordance with one of the following documents: (1) SNT-TC-1A (2016 Edition),1 Personnel Qualification and Certification in Nondestructive Testing, as amended by Mandatory Appendix III; or 1 ð21Þ ARTICLE 1 ASME BPVC.V-2021 (2) ANSI/ASNT CP-189 (2020 Edition),1 ASNT Standard for Qualification and Certification of Nondestructive Testing Personnel, as amended by Mandatory Appendix IV (f) National or international central certification programs, such as ISO 9712–based programs, may alternatively be used to fulfill the written practice requirements of (e) for training, experience, general examination, basic examination, and method examination, as applicable. (g) In addition to the requirements described in (e) or (f) above, if the techniques of computed radiography (CR), digital radiography (DR), phased-array ultrasonic (PAUT), ultrasonic time-of-flight diffraction (TOFD), or ultrasonic full matrix capture (FMC) are to be used, the training, experience, and examination requirements found in Article 1, Mandatory Appendix II shall also be included in the employer’s written practice for each technique as applicable. (h) Alternatively, performance-based qualification programs, in accordance with ASME ANDE-1-2015, ASME Nondestructive Examination and Quality Control Central Qualification and Certification Program, may be used for training, experience, examination, and certification activities as specified in the written practice. (i) When the referencing Code Section does not specify qualifications or does not reference directly Article 1 of this Section, qualification may simply involve a demonstration to show that the personnel performing the nondestructive examinations are competent to do so in ac cordance with the organizatio n’s established procedures. (j) The user of this Article is responsible for the qualification and certification of NDE Personnel in accordance with the requirements of this Article. The organization’s2 Quality Program shall stipulate how this is to be accomplished. Qualifications in accordance with a prior edition of SNT-TC-1A, or CP-189 are valid until recertification. Recertification or new certification shall be in accordance with the edition of SNT-TC-1A or CP-189 specified in (e) above. When any of the techniques included in (g) above are used, the additional requirements of that paragraph shall also apply. (k) Limited certification of nondestructive examination personnel who do not perform all of the operations of a nondestructive method that consists of more than one operation, or who perform nondestructive examinations of limited scope, may be based on fewer hours of training and experience than recommended in SNT-TC-1A or CP-189. Any limitations or restrictions placed upon a person’s certification shall be described in the written practice and on the certification. (l) Either U.S. Customary Units or SI Units may be used for compliance with all requirements of this edition, but one system shall be used consistently throughout for all phases of construction. (1) Either the U.S. Customary Units or SI Units that are listed in Section V Mandatory Appendix II (in the rear of Section V and listed in other Code books) are identified in the text, or are identified in the nomenclature for equations shall be used consistently for all phases of construction (e.g., materials, design, fabrication, and reports). Since values in the two systems are not exact equivalents, each system shall be used independently of the other without mixing U.S. Customary Units and SI Units. (2) When SI Units are selected, U.S. Customary values in referenced specifications that do not contain SI Units shall be converted to SI values to at least three significant figures for use in calculations and other aspects of construction. T-130 EQUIPMENT It is the responsibility of the Code User to ensure that the examination equipment being used conforms to the requirements of this Code Section. T-150 PROCEDURE (a) When required by the referencing Code Section, all nondestructive examinations performed under this Code Section shall be performed following a written procedure. A procedure demonstration shall be performed to the satisfaction of the Inspector. When required by the referencing Code Section, a personnel demonstration may be used to verify the ability of the examiner to apply the examination procedure. The examination procedure shall comply with the applicable requirements of this Section for the particular examination method. Written procedures shall be made available to the Inspector on request. At least one copy of each procedure shall be readily available to the Nondestructive Examination Personnel for their reference and use. (b) The nondestructive examination methods and techniques included in this Section are applicable to most geometric configurations and materials encountered in fabrication under normal conditions. Whenever special configurations or materials require modified methods and techniques, the organization shall develop special procedures which are equivalent or superior to the methods and techniques described in this Code Section, and which are capable of producing interpretable examination results under the special conditions. Such special procedures may be modifications or combinations of methods described or referenced in this Code Section. A procedure demonstration shall be performed to verify the technique is capable of detecting discontinuities under the special conditions equal to the capabilities of the method when used under more general conditions. These special procedures shall be submitted to the 2 ASME BPVC.V-2021 Inspector for acceptance when required by the referencing Code Section, and shall be adopted as part of the Manufacturer’s quality control program. (c) When a referencing Code Section requires an examination to be performed in accordance with the requirements of this Section, it shall be the responsibility of the organization to establish nondestructive examination procedures and personnel qualification and certification procedures conforming to the referenced requirements. (d) When qualification of the written examination procedure is required by the referencing Code Section, a qualification demonstration shall be performed prior to acceptance of production examinations. The qualification demonstration shall be performed (1) under the control and supervision of a Level III Examiner who is qualified and certified for performing the examination method and technique specified by the procedure, and shall be witnessed by the Inspector. The supervising Level III may be an employee of the qualifying organization or a subcontractor organization. (2) on a minimum of one test specimen having flaws whose size, location, orientation, quantity, and characterization have been determined prior to the demonstration and are known only by the supervising Level III Examiner. (-a) The maximum acceptable flaw size, required flaw orientation, and minimum number of flaws shall be as specified by the referencing Code Section. (-b) Natural flaws are preferred over artificial flaws whenever possible. (3) by a Level II or Level III Examiner (other than the supervising Level III) who is qualified and certified to perform the examination method and technique specified by the written procedure. The procedure shall be considered qualified when the supervising Level III and the Inspector are satisfied that indications produced by the demonstrated procedure effectively reveal the size, location, orientation, quantity, and characterization of the flaws known to be present in the examined test specimen. The qualification demonstration shall be documented as required by the referencing Code Section and by this Section, as set forth in the applicable Article for the examination method and the applicable Appendix for the specified examination technique. The qualification document shall be annotated to indicate qualification of the written procedure, and identify the examined test specimen. The name and/or identity and signature of the supervising Level III and the witnessing Inspector shall be added to indicate their acceptance of the procedure qualification. ð21Þ T-160 ARTICLE 1 (b) When special procedures are developed [see T-150(b)], the Code User shall specify what calibration is necessary, when calibration is required. T-170 EXAMINATIONS AND INSPECTIONS (a) The Inspector concerned with the fabrication of the vessel or pressure part shall have the duty of verifying to the Inspector's satisfaction that all examinations required by the referencing Code Section have been made to the requirements of this Section and the referencing document(s). The Inspector shall have the right to witness any of these examinations to the extent stated in the referencing document(s). Throughout this Section of the Code, the word Inspector shall be as defined and qualified as required by the referencing Code Section or referencing document(s). (b) The special distinction established in the various Code Sections between inspection and examination and the personnel performing them is also adopted in this Code Section. In other words, the term inspection applies to the functions performed by the Inspector, but the term examination applies to those quality control functions performed by personnel employed by the organization. One area of occasional deviation from these distinctions exists. In the ASTM Standard Methods and Recommended Practices incorporated in this Section of the Code by reference or by reproduction in Subsection B, the words inspection or Inspector, which frequently occur in the text or titles of the referenced ASTM documents, may actually describe what the Code calls examination or examiner. This situation exists because ASTM has no occasion to be concerned with the distinctions which the Code makes between inspection and examination, since ASTM activities and documents do not involve the Inspector described in the Code Sections. However, no attempt has been made to edit the ASTM documents to conform with Code usage; this should cause no difficulty if the users of this Section recognize that the terms inspection, testing, and examination in the ASTM documents referenced in Subsection B do not describe duties of the Inspector but rather describe the things to be done by the organization’s examination personnel. T-180 EVALUATION The acceptance criteria for the NDE methods in this Section shall be as stated in the referencing Code Section, and where provided in the Articles of this Section. Acceptance criteria in the referencing Code Section shall take precedence. CALIBRATION (a) The organization shall assure that all equipment calibrations required by Subsection A and/or Subsection B are performed. 3 ARTICLE 1 T-190 ASME BPVC.V-2021 RECORDS/DOCUMENTATION T-150(a) or T-150(b) are not specified by the referencing Code Section, the following information shall be recorded as a minimum: (1) name of organization responsible for preparation and approval of the examination procedure (2) examination method applied (3) procedure number or designation (4) number and date of most recent revision (5) date of the demonstration (6) name and/or identity and certification level (if applicable) of personnel performing demonstration (d) Retention of examination records and related documentation (e.g., radiographs and review forms, ultrasonic scan files, etc.) shall be as specified by the referencing Code Section. (e) Digital images and reviewing software shall be retained under an appropriate record retention system that is capable of securely storing and retrieving data for the time period specified by the referencing Code Section. (a) Documentation and records shall be prepared as specified by the referencing Code Section and the applicable requirements of this Section. Examination records shall include the following information as a minimum: (1) date of the examination (2) name and/or identity and certification level (if applicable) for personnel performing the examination (3) identification of the weld, part, or component examined including weld number, serial number, or other identifier (4) examination method, technique, procedure identification, and revision (5) results of the examination (b) Personnel qualification and procedure performance demonstrations performed in compliance with the requirements of T-150(a) or T-150(b) shall be documented as specified by the referencing Code Section. (c) When documentation requirements for personnel qualification and procedure performance demonstrations performed in compliance with the requirements of 4 ASME BPVC.V-2021 ARTICLE 2 ARTICLE 2 RADIOGRAPHIC EXAMINATION T-210 SCOPE T-222.2 Welds. The weld ripples or weld surface irregularities on both the inside (where accessible) and outside shall be removed by any suitable process to such a degree that the images of surface irregularities cannot mask or be confused with the image of any discontinuity on the resulting radiograph. The finished surface of all butt-welded joints may be flush with the base material or may have reasonably uniform crowns, with reinforcement not to exceed that specified in the referencing Code Section. The radiographic method described in this Article for examination of materials including castings and welds shall be used together with Article 1, General Requirements. Definitions of terms used in this Article are in Article 1, Mandatory Appendix I, I-121.1, RT — Radiography. Certain product-specific, technique-specific, and application-specific requirements are also given in other Mandatory Appendices of this Article, as listed in the table of contents. These additional requirements shall also be complied with when an Appendix is applicable to the r a d i o g r a p h i c o r r a d i o s c opic examination being conducted. T-220 GENERAL REQUIREMENTS T-221 PROCEDURE REQUIREMENTS T-223 A lead symbol “B,” with minimum dimensions of 7/16 in. (11 mm) in height and 1/16 in. (1.5 mm) in thickness, shall be attached to the back of each film holder during each exposure to determine if backscatter radiation is exposing the film. The lead symbol “B” shall be placed in a location so that it would appear within an area on the radiograph that meets the requirements of T-282, VIII-288, or IX-288, as applicable. T-221.1 Written Procedure. Radiographic examination shall be performed in accordance with a written procedure. Each procedure shall include at least the following information, as applicable: (a) material type and thickness range (b) isotope or maximum X-ray voltage used (c) source-to-object distance (D in T-274.1) (d) distance from source side of object to film (d in T-274.1) (e) source size (F in T-274.1) (f) film brand and designation (g) screens used T-224 SYSTEM OF IDENTIFICATION A system shall be used to produce on each radiograph an identification that is traceable to the item being radiographed and that is permanent for the required retention period of the radiograph. This information shall include the contract, component, weld number, or part number, as appropriate. In addition, the Manufacturer’s symbol or name and the date of the radiograph shall be included with the identification information on each radiograph. An NDE subcontractor’s name or symbol may also be used together with that of the Manufacturer. This identification system does not necessarily require that the information appear as radiographic images. In any case, this information shall not obscure the area of interest. T-221.2 Procedure Demonstration. Demonstration of the density and image quality indicator (IQI) image requirements of the written procedure on production or technique radiographs shall be considered satisfactory evidence of compliance with that procedure. T-225 T-222 BACKSCATTER RADIATION SURFACE PREPARATION MONITORING DENSITY LIMITATIONS OF RADIOGRAPHS Either a densitometer or step wedge comparison film shall be used for judging film density. T-222.1 Materials Including Castings. Surfaces shall satisfy the requirements of the applicable materials specification or referencing Code Section, with additional conditioning, if necessary, by any suitable process to such a degree that the images of surface irregularities cannot mask or be confused with the image of any discontinuity on the resulting radiograph. T-226 EXTENT OF EXAMINATION The extent of radiographic examination shall be as specified by the referencing Code Section. 41 ð21Þ ARTICLE 2 T-230 T-231 ASME BPVC.V-2021 EQUIPMENT AND MATERIALS Table T-233.2 Wire IQI Designation, Wire Diameter, and Wire Identity FILM T-231.1 Selection. Radiographs shall be made using industrial radiographic film. ð21Þ Set A Wire Diameter, in. (mm) T-231.2 Processing. Standard Guide for Controlling the Quality of Industrial Radiographic Film Processing, SE-999, or Sections 23 through 26 of Standard Guide for Radiographic Examination Using Industrial Radiographic Film, SE-94/SE-94M, may be used as a guide for processing film, except that Section 8.1 of SE-999 is not required. T-232 0.0032 0.004 0.005 0.0063 0.008 0.010 INTENSIFYING SCREENS Wire Identity Wire Diameter, in. (mm) Wire Identity 1 2 3 4 5 6 0.010 (0.25) 0.013 (0.33) 0.016 (0.41) 0.020 (0.51) 0.025 (0.64) 0.032 (0.81) 6 7 8 9 10 11 Wire Identity Wire Diameter, in. (mm) Wire Identity 0.100 (2.54) 0.126 (3.20) 0.160 (4.06) 0.200 (5.08) 0.250 (6.35) 0.320 (8.13) 16 17 18 19 20 21 (0.08) (0.10) (0.13) (0.16) (0.20) (0.25) Set C Intensifying screens may be used when performing radiographic examination in accordance with this Article. T-233 Set B Wire Diameter, in. (mm) 0.032 (0.81) 0.040 (1.02) 0.050 (1.27) 0.063 (1.60) 0.080 (2.03) 0.100 (2.54) IMAGE QUALITY INDICATOR (IQI) DESIGN T-233.1 Standard IQI Design. IQIs shall be either the hole type or the wire type. Hole-type IQIs shall be manufactured and identified in accordance with the requirements or alternates allowed in SE-1025. Wire-type IQIs shall be manufactured and identified in accordance with the requirements or alternates allowed in SE-747, except that the largest wire number or the identity number may be omitted. ASME standard IQIs shall consist of those in Table T-233.1 for hole type and those in Table T-233.2 for wire type. Set D 11 12 13 14 15 16 Table T-233.1 Hole-Type IQI Designation, Thickness, and Hole Diameters IQI Designation 5 7 10 12 15 17 20 25 30 35 40 45 50 60 70 80 100 120 140 160 200 240 280 IQI Thickness, in. (mm) 0.005 0.0075 0.010 0.0125 0.015 0.0175 0.020 0.025 0.030 0.035 0.040 0.045 0.050 0.060 0.070 0.080 0.100 0.120 0.140 0.160 0.200 0.240 0.280 (0.13) (0.19) (0.25) (0.32) (0.38) (0.44) (0.51) (0.64) (0.76) (0.89) (1.02) (1.14) (1.27) (1.52) (1.78) (2.03) (2.54) (3.05) (3.56) (4.06) (5.08) (6.10) (7.11) 1T Hole Diameter, in. (mm) 0.010 0.010 0.010 0.0125 0.015 0.0175 0.020 0.025 0.030 0.035 0.040 0.045 0.050 0.060 0.070 0.080 0.100 0.120 0.140 0.160 0.200 0.240 0.280 42 (0.25) (0.25) (0.25) (0.32) (0.38) (0.44) (0.51) (0.64) (0.76) (0.89) (1.02) (1.14) (1.27) (1.52) (1.78) (2.03) (2.54) (3.05) (3.56) (4.06) (5.08) (6.10) (7.11) 2T Hole Diameter, in. (mm) 0.020 0.020 0.020 0.025 0.030 0.035 0.040 0.050 0.060 0.070 0.080 0.090 0.100 0.120 0.140 0.160 0.200 0.240 0.280 0.320 0.400 0.480 0.560 (0.51) (0.51) (0.51) (0.64) (0.76) (0.89) (1.02) (1.27) (1.52) (1.78) (2.03) (2.29) (2.54) (3.05) (3.56) (4.06) (5.08) (6.10) (7.11) (8.13) (10.16) (12.19) (14.22) 4T Hole Diameter, in. (mm) 0.040 (1.02) 0.040 (1.02) 0.040 (1.02) 0.050 (1.27) 0.060 (1.52) 0.070 (1.78) 0.080 (2.03) 0.100 (2.54) 0.120 (3.05) 0.140 (3.56) 0.160 (4.06) 0.180 (4.57) 0.200 (5.08) 0.240 (6.10) 0.280 (7.11) 0.320 (8.13) 0.400 (10.16) 0.480 (12.19) 0.560 (14.22) 0.640 (16.26) … … … ASME BPVC.V-2021 T-233.2 Alternative IQI Design. IQIs designed and manufactured in accordance with other national or international standards may be used provided the requirements of either (a) or (b) below, and the material requirements of T-276.1 are met. (a) Hole-Type IQIs. The calculated Equivalent IQI Sensitivity (EPS), per SE-1025, Appendix X1, is equal to or better than the required standard hole-type IQI. (b) Wire-Type IQIs. The alternative wire IQI essential wire diameter is equal to or less than the required standard IQI essential wire. T-234 by comparison with a national standard step tablet unless, prior to first use, it was maintained in the original light-tight and waterproof sealed package as supplied by the manufacturer. Step wedge calibration films may be used without verification for one year upon opening, provided it is within the manufacturer’s stated shelf life. (b) The densitometer manufacturer’s step-by-step instructions for the operation of the densitometer shall be followed. (c) The density steps closest to 1.0, 2.0, 3.0, and 4.0 on the national standard step tablet or step wedge calibration film shall be read. (d) The densitometer is acceptable if the density readings do not vary by more than ±0.05 density units from the actual density stated on the national standard step tablet or step wedge calibration film. FACILITIES FOR VIEWING OF RADIOGRAPHS Viewing facilities shall provide subdued background lighting of an intensity that will not cause reflections, shadows, or glare on the radiograph that interfere with the interpretation process. Equipment used to view radiographs for interpretation shall provide a variable light source sufficient for the essential IQI hole or designated wire to be visible for the specified density range. The viewing conditions shall be such that light from around the outer edge of the radiograph or coming through lowdensity portions of the radiograph does not interfere with interpretation. T-260 T-261 T-262.2 Step Wedge Comparison Films. Step wedge comparison films shall be verified prior to first use, unless performed by the manufacturer, as follows: (a) The density of the steps on a step wedge comparison film shall be verified by a calibrated densitometer. (b) The step wedge comparison film is acceptable if the density readings do not vary by more than ±0.1 density units from the density stated on the step wedge comparison film. T-262.3 Periodic Verification. ð21Þ (a) Densitometers. Periodic calibration verification checks shall be performed as described in T-262.1 at the beginning of each shift, after 8 hr of continuous use, or after change of apertures, whichever comes first. (b) Step Wedge Comparison Films. Verification checks shall be performed annually per T-262.2. CALIBRATION SOURCE SIZE T-261.1 Verification of Source Size. The equipment manufacturer’s or supplier’s publications, such as technical manuals, decay curves, or written statements documenting the actual or maximum source size or focal spot, shall be acceptable as source size verification. T-262.4 Documentation. (a) Densitometers. Densitometer calibrations required by T-262.1 shall be documented, but the actual readings for each step do not have to be recorded. Periodic densitometer verification checks required by T-262.3(a) do not have to be documented. (b) Step Wedge Calibration Films. Step wedge calibration film verifications required by T-262.1(a) shall be documented, but the actual readings for each step do not have to be recorded. (c) Step Wedge Comparison Films. Step wedge comparison film verifications required by T-262.2 and T-262.3(b) shall be documented, but the actual readings for each step do not have to be recorded. T-261.2 Determination of Source Size. When manufacturer’s or supplier’s publications are not available, source size may be determined as follows: (a) X-Ray Machines. For X-ray machines operating at 1,000 kV and less, the focal spot size may be determined in accordance with SE-1165, Standard Test Method for Measurement of Focal Spots of Industrial X-Ray Tubes by Pinhole Imaging. (b) Iridium-192 Sources. For Iridium-192, the source size may be determined in accordance with SE-1114, Standard Test Method for Determining the Focal Size of Iridium-192 Industrial Radiographic Sources. T-262 ARTICLE 2 DENSITOMETER AND STEP WEDGE COMPARISON FILM T-262.1 Densitometers. Densitometers shall be calibrated at least every 3 months during use as follows: (a) A national standard step tablet or a step wedge calibration film, traceable to a national standard step tablet and having at least five steps with neutral densities from at least 1.0 through 4.0, shall be used. The step wedge calibration film shall have been verified within the last year T-270 T-271 EXAMINATION RADIOGRAPHIC TECHNIQUE3 A single-wall exposure technique shall be used for radiography whenever practical. When it is not practical to use a single-wall technique, a double-wall technique 43 ARTICLE 2 ASME BPVC.V-2021 T-274 shall be used. An adequate number of exposures shall be made to demonstrate that the required coverage has been obtained. T-274.1 Geometric Unsharpness Determination. ð21Þ Geometric unsharpness of the radiograph shall be determined in accordance with: T-271.1 Single-Wall Technique. In the single-wall technique, the radiation passes through only one wall of the weld (material), which is viewed for acceptance on the radiograph. where T-271.2 Double-Wall Technique. When it is not practical to use a single-wall technique, one of the following double-wall techniques shall be used. D = distance from source of radiation to weld or object being radiographed d = distance from source side of weld or object being radiographed to the film F = source size: the maximum projected dimension of the radiating source (or effective focal spot) in the plane perpendicular to the distance D from the weld or object being radiographed U g = geometric unsharpness (a) Single-Wall Viewing. For materials and for welds in components, a technique may be used in which the radiation passes through two walls and only the weld (material) on the film-side wall is viewed for acceptance on the radiograph. When complete coverage is required for circumferential welds (materials), a minimum of three exposures taken 120 deg to each other shall be made. D and d shall be determined at the approximate center of the area of interest. (b) Double-Wall Viewing. For materials and for welds in components 31/2 in. (89 mm) or less in nominal outside diameter, a technique may be used in which the radiation passes through two walls and the weld (material) in both walls is viewed for acceptance on the same radiograph. For double-wall viewing, only a source-side IQI shall be used. NOTE: Alternatively, a nomograph as shown in Standard Guide for Radiographic Examination Using Industrial Radiographic Film, SE-94/SE-94M, may be used. T-274.2 Geometric Unsharpness Limitations. Recommended maximum values for geometric unsharpness are as follows: (1) For welds, the radiation beam may be offset from the plane of the weld at an angle sufficient to separate the images of the source-side and film-side portions of the weld so that there is no overlap of the areas to be interpreted. When complete coverage is required, a minimum of two exposures taken 90 deg to each other shall be made for each joint. Material Thickness, in. (mm) Under 2 (50) 2 through 3 (50–75) Over 3 through 4 (75–100) Greater than 4 (100) (2) As an alternative, the weld may be radiographed with the radiation beam positioned so that the images of both walls are superimposed. When complete coverage is required, a minimum of three exposures taken at either 60 deg or 120 deg to each other shall be made for each joint. T-275 0.020 0.030 0.040 0.070 (0.51) (0.76) (1.02) (1.78) LOCATION MARKERS Location markers (see Figure T-275), which shall appear as radiographic images on the radiograph, shall be placed on the part, not on the exposure holder/cassette. Their locations shall be permanently marked on the surface of the part being radiographed when permitted, or on a map, in a manner permitting the area on a radiograph to be accurately traceable to its location on the part, for the required retention period of the radiograph. In addition, their locations do not limit the area of interest or the area to be interpreted. Evidence shall also be provided on the radiograph that the required coverage of the region being examined has been obtained. Location markers shall be placed as follows. RADIATION ENERGY The radiation energy employed for any radiographic technique shall achieve the density and IQI image requirements of this Article. T-273 Ug Maximum, in. (mm) NOTE: Material thickness is the thickness on which the IQI is based. (3) Additional exposures shall be made if the required radiographic coverage cannot be obtained using the minimum number of exposures indicated in (1) or (2) above. T-272 GEOMETRIC UNSHARPNESS T-275.1 Single-Wall Viewing. (a) Source-Side Markers. Location markers shall be placed on the source side when radiographing the following: (1) flat components or longitudinal joints in cylindrical or conical components; DIRECTION OF RADIATION The direction of the central beam of radiation should be centered on the area of interest whenever practical. 44 ð21Þ Figure T-275 Location Marker Sketches Source side acceptable Film side unacceptable Source side acceptable Film side unacceptable Source side acceptable Film side unacceptable Curved components with radiation source to film distance less than radius of component [See T-275.1(a)(2)] (b) Curved components with convex surface towards radiation source [See T-275.1(a)(3)] (c) 45 Either side location marker is acceptable D ASME BPVC.V-2021 Flat component or longitudinal seam [See T-275.1(a)(1)] [See sketch (e) for alternate] (a) t Source side unacceptable x Film side acceptable Curved components with radiation source to film distance greater than radius of curvature [See T-275.1(b)(1)] (d) Radiation source — Location marker — Component center — x Source side marker alternate Flat component or logitudinal seam x = (t / D) (Mf / 2) x = additional required coverage beyond film side location marker t = component thickness Mf = film side location marker interval D = source to component distance [See T-275.1(b)(2)] (e) Curved components with radiation source at center curvature [See T-275.1(c)] (f) ARTICLE 2 LEGEND: Mf ARTICLE 2 ASME BPVC.V-2021 referencing Code Section. Physical measurement of the actual weld reinforcements is not required. Backing rings or strips shall not be considered as part of the thickness in IQI selection. (2) curved or spherical components whose concave side is toward the source and when the “source-tomaterial” distance is less than the inside radius of the component; (3) curved or spherical components whose convex side is toward the source. (b) Film-Side Markers (1) Location markers shall be placed on the film side when radiographing either curved or spherical components whose concave side is toward the source and when the “source-to-material” distance is greater than the inside radius. (2) As an alternative to source-side placement in T-275.1(a)(1), location markers may be placed on the film side when the radiograph shows coverage beyond the location markers to the extent demonstrated by Figure T-275, sketch (e), and when this alternate is documented in accordance with T-291. (c) Either Side Markers. Location markers may be placed on either the source side or film side when radiographing either curved or spherical components whose concave side is toward the source and the “source-tomaterial” distance equals the inside radius of the component. (b) Welds Without Reinforcements. The thickness on which the IQI is based is the nominal single-wall material thickness. Backing rings or strips shall not be considered as part of the thickness in IQI selection. (c) Actual Values. With regard to (a) and (b) above, when the actual material/weld thickness is measured, IQI selection may be based on these known values. T-276.3 Welds Joining Dissimilar Materials or Welds With Dissimilar Filler Metal. When the weld metal is of an alloy group or grade that has a radiation attenuation that differs from the base material, the IQI material selection shall be based on the weld metal and be in accordance with T-276.1. When the density limits of T-282.2 cannot be met with one IQI, and the exceptional density area(s) is at the interface of the weld metal and the base metal, the material selection for the additional IQIs shall be based on the base material and be in accordance with T-276.1. T-275.2 Double-Wall Viewing. For double-wall viewing, at least one location marker shall be placed adjacent to the weld (or on the material in the area of interest) for each radiograph. T-277 T-277.1 T-275.3 Mapping the Placement of Location Markers. When inaccessibility or other limitations prevent the placement of markers as stipulated in T-275.1 and T-275.2, a dimensioned map of the actual marker placement shall accompany the radiographs to show that full coverage has been obtained. T-276 Placement of IQIs. (a) Source-Side IQI(s). The IQI(s) shall be placed on the source side of the part being examined, except for the condition described in (b). When, due to part or weld configuration or size, it is not practical to place the IQI(s) on the part or weld, the IQI(s) may be placed on a separate block. Separate blocks shall be made of the same or radiographically similar materials (as defined in SE-1025 for hole type or SE-747 for wire type) and may be used to facilitate IQI positioning. There is no restriction on the separate block thickness, provided the IQI/area-of-interest density tolerance requirements of T-282.2 are met. IQI SELECTION T-276.1 Material. IQIs shall be selected from either the same alloy material group or grade as identified in SE-1025 for hole type or SE-747 for wire type, or from an alloy material group or grade with less radiation absorption than the material being radiographed. ð21Þ USE OF IQIS TO MONITOR RADIOGRAPHIC EXAMINATION (1) The IQI on the source side of the separate block shall be placed no closer to the film than the source side of the part being radiographed. T-276.2 Size. The designated hole IQI or essential wire shall be as specified in Table T-276. A thinner or thicker hole-type IQI may be substituted for any section thickness listed in Table T-276, provided an equivalent IQI sensitivity is maintained. See T-283.2. For wire-type IQIs, a smaller-diameter wire may be substituted for the essential wire required for any section thickness listed in Table T-276. (a) Welds With Reinforcements. The thickness on which the IQI is based is the nominal single-wall material thickness plus the weld reinforcement thickness estimated to be present on both sides of the weld (I.D. and O.D.). The values used for the estimated weld reinforcement thicknesses shall be representative of the weld conditions and shall not exceed the maximums permitted by the (2) The separate block shall be placed as close as possible to the part being radiographed. (3) When hole-type IQIs are used, the block dimensions shall exceed the IQI dimensions such that the outline of at least three sides of the IQI image shall be visible on the radiograph. (b) Film-Side IQI(s). Where inaccessibility prevents hand placing the IQI(s) on the source side, the IQI(s) shall be placed on the film side in contact with the part being examined. A lead letter “F” shall be placed adjacent to or on the IQI(s), but shall not mask the essential hole where hole IQIs are used. 46 ð21Þ ASME BPVC.V-2021 ARTICLE 2 ð21Þ Table T-276 IQI Selection IQI Source Side Nominal Single-Wall Material Thickness Range, in. (mm) ≤0.25 (≤6.4) >0.25 through 0.375 (>6.4 through 9.5) >0.375 through 0.50 (>9.5 through 12.7) >0.50 through 0.75 (>12.7 through 19.0) >0.75 through 1.00 (>19.0 through 25.4) >1.00 through 1.50 (>25.4 through 38.1) >1.50 through 2.00 (>38.1 through 50.8) >2.00 through 2.50 (>50.8 through 63.5) >2.50 through 4.00 (>63.5 through 101.6) >4.00 through 6.00 (>101.6 through 152.4) >6.00 through 8.00 (>152.4 through 203.2) >8.00 through 10.00 (>203.2 through 254.0) >10.00 through 12.00 (>254.0 through 304.8) >12.00 through 16.00 (>304.8 through 406.4) >16.00 through 20.00 (>406.4 through 508.0) Film Side Hole-Type Designation Essential Hole Wire-Type Essential Wire Hole-Type Designation Essential Hole Wire-Type Essential Wire 12 15 17 20 25 30 35 40 50 60 80 100 120 160 200 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 5 6 7 8 9 10 11 12 13 14 16 17 18 20 21 10 12 15 17 20 25 30 35 40 50 60 80 100 120 160 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 2T 4 5 6 7 8 9 10 11 12 13 14 16 17 18 20 (c) IQI Placement for Welds — Hole IQIs. The IQI(s) may be placed adjacent to or on the weld. The identification number(s) and, when used, the lead letter “F,” shall not be in the area of interest, except when geometric configuration makes it impractical. (d) IQI Placement for Welds — Wire IQIs. The IQI(s) shall be placed on the weld so that the lengths of the wires are transverse to the longitudinal axis of the weld. The IQI identification and, when used, the lead letter “F,” shall not be in the area of interest, except when geometric configuration makes it impractical. (e) IQI Placement for Materials Other Than Welds. The IQI(s) with the IQI identification and, when used, the lead letter “F,” may be placed in the area of interest. (-b) When a section or sections of the circumference, where the length between the ends of the outermost sections span 240 or more deg, is radiographed using one or more film holders. Additional film locations may be required to obtain necessary IQI spacing. (2) For cylindrical components where the source is placed on the axis of the component for a single exposure, at least three IQIs, with one placed at each end of the span of the circumference radiographed and one in the approximate center of the span, are required under the following conditions: (-a) When a section of the circumference, the length of which is greater than 120 deg and less than 240 deg, is radiographed using just one film holder, or; (-b) When a section or sections of the circumference, where the length between the ends of the outermost sections span less than 240 deg, is radiographed using more than one film holder. (3) In (1) and (2) above, where sections of longitudinal welds adjoining the circumferential weld are radiographed simultaneously with the circumferential weld, an additional IQI shall be placed on each longitudinal weld at the end of the section most remote from the junction with the circumferential weld being radiographed. (4) For spherical components where the source is placed at the center of the component for a single exposure, at least three IQIs, spaced approximately 120 deg apart, are required under the following conditions: (-a) When a complete circumference is radiographed using one or more film holders, or; T-277.2 Number of IQIs. When one or more film holders are used for an exposure, at least one IQI image shall appear on each radiograph except as outlined in (b) below. (a) Multiple IQIs. If the requirements of T-282 are met by using more than one IQI, one shall be representative of the lightest area of interest and the other the darkest area of interest; the intervening densities on the radiograph shall be considered as having acceptable density. (b) Special Cases4 (1) For cylindrical components where the source is placed on the axis of the component for a single exposure, at least three IQIs, spaced approximately 120 deg apart, are required under the following conditions: (-a) When the complete circumference is radiographed using one or more film holders, or; 47 ARTICLE 2 ASME BPVC.V-2021 (-b) When a section or sections of a circumference, where the length between the ends of the outermost sections span 240 or more deg, is radiographed using one or more film holders. Additional film locations may be required to obtain necessary IQI spacing. (5) For spherical components where the source is placed at the center of the component for a single exposure, at least three IQIs, with one placed at each end of the span of the circumference radiographed and one in the approximate center of the span, are required under the following conditions: (-a) When a section of a circumference, the length of which is greater than 120 deg and less than 240 deg, is radiographed using just one film holder, or; (-b) When a section or sections of a circumference, where the length between the ends of the outermost sections span less than 240 deg is radiographed using more than one film holder. (6) In (4) and (5) above, where other welds are radiographed simultaneously with the circumferential weld, one additional IQI shall be placed on each other weld. (7) For segments of a flat or curved (i.e., ellipsoidal, torispherical, toriconical, elliptical, etc.) component where the source is placed perpendicular to the center of a length of weld for a single exposure when using more than three film holders, at least three IQIs, one placed at each end of the radiographed span and one in the approximate center of the span, are required. (8) When an array of components in a circle is radiographed, at least one IQI shall show on each component image. (9) In order to maintain the continuity of records involving subsequent exposures, all radiographs exhibiting IQIs that qualify the techniques permitted in accordance with (1) through (7) above shall be retained. (b) processing defects such as streaks, watermarks, or chemical stains; (c) scratches, finger marks, crimps, dirtiness, static marks, smudges, or tears; (d) false indications due to defective screens. T-282 T-282.1 Density Limitations. The transmitted film density through the radiographic image of the body of the designated hole-type IQI adjacent to the essential hole or adjacent to the essential wire of a wire-type IQI and the area of interest shall be 1.8 minimum for single film viewing for radiographs made with an X-ray source and 2.0 minimum for radiographs made with a gamma ray source. For composite viewing of multiple film exposures, each film of the composite set shall have a minimum density of 1.3. The maximum density shall be 4.0 for either single or composite viewing. A tolerance of 0.05 in density is allowed for variations between densitometer readings. T-282.2 Density Variation. (a) The density of the radiograph anywhere through the area of interest shall not (1) vary by more than minus 15% or plus 30% from the density through the body of the designated hole-type IQI adjacent to the essential hole or adjacent to the essential wire of a wire-type IQI, and (2) exceed the minimum/maximum allowable density ranges specified in T-282.1 When calculating the allowable variation in density, the calculation may be rounded to the nearest 0.1 within the range specified in T-282.1. (b) When the requirements of (a) above are not met, then an additional IQI shall be used for each exceptional area or areas and the radiograph retaken. (c) When shims are used with hole-type IQIs, the plus 30% density restriction of (a) above may be exceeded, and the minimum density requirements of T-282.1 do not apply for the IQI, provided the required IQI sensitivity of T-283.1 is met. T-277.3 Shims Under Hole-Type IQIs. For welds, a shim of material radiographically similar to the weld metal shall be placed between the part and the IQI, if needed, so that the radiographic density throughout the area of interest is no more than minus 15% from (lighter than) the radiographic density through the designated IQI adjacent to the essential hole. The shim dimensions shall exceed the IQI dimensions such that the outline of at least three sides of the IQI image shall be visible in the radiograph. T-280 T-281 RADIOGRAPHIC DENSITY T-283 IQI SENSITIVITY T-283.1 Required Sensitivity. Radiography shall be ð21Þ performed with a technique of sufficient sensitivity to display the designated hole-type IQI image (including applicable material group identification notches) and the essential hole, or the essential wire of a wire-type IQI. The radiographs shall also display the IQI identifying numbers and letters. If the designated hole-type IQI image (including applicable material group identification notches) and essential hole, or essential wire of a wiretype IQI, do not show on any film in a multiple film technique, but do show in composite film viewing, interpretation shall be permitted only by composite film viewing. EVALUATION QUALITY OF RADIOGRAPHS All radiographs shall be free from mechanical, chemical, or other blemishes to the extent that they do not mask and are not confused with the image of any discontinuity in the area of interest of the object being radiographed. Such blemishes include, but are not limited to: (a) fogging; 48 ASME BPVC.V-2021 the radiograph review form documentation shall accompany the radiographs. Acceptance shall be completed prior to presentation of the radiographs and accompanying documentation to the Inspector. For wire-type IQIs, the essential wire shall be visible within the area of interest representing the thickness used for determining the essential wire, inclusive of the allowable density variations described in T-282.2. T-283.2 Equivalent Hole-Type IQI Sensitivity. A thinner or thicker hole-type IQI than the designated IQI may be substituted, provided an equivalent or better IQI sensitivity, as listed in Table T-283, is achieved and all other requirements for radiography are met. Equivalent IQI sensitivity is shown in any row of Table T-283 which contains the designated IQI and hole. Better IQI sensitivity is shown in any row of Table T-283 which is above the equivalent sensitivity row. If the designated IQI and hole are not represented in the table, the next thinner IQI row from Table T-283 may be used to establish equivalent IQI sensitivity. T-284 T-290 T-291 EXCESSIVE BACKSCATTER EVALUATION BY MANUFACTURER The Manufacturer shall be responsible for the review, interpretation, evaluation, and acceptance of the completed radiographs to assure compliance with the requirements of Article 2 and the referencing Code Section. As an aid to the review and evaluation, the radiographic technique documentation required by T-291 shall be completed prior to the evaluation. The radiograph review form required by T-292 shall be completed during the evaluation. The radiographic technique details and T-292 10 12 15 17 20 25 30 35 40 50 60 80 100 120 160 200 Equivalent Hole-Type Designations 1T Hole 4T Hole 15 17 20 25 30 35 40 50 60 70 80 120 140 160 240 280 5 7 10 12 15 17 20 25 30 35 40 60 70 80 120 140 RADIOGRAPHIC TECHNIQUE DOCUMENTATION DETAILS RADIOGRAPH REVIEW FORM The Manufacturer shall be responsible for the preparation of a radiograph review form. As a minimum, the following information shall be provided. (a) a listing of each radiograph location (b) the information required in T-291, by inclusion of the information on the review form or by reference to an attached radiographic technique details sheet (c) evaluation and disposition of the material(s) or weld(s) examined (d) identification (name) of the Manufacturer’s representative who performed the final acceptance of the radiographs (e) date of Manufacturer’s evaluation Table T-283 Equivalent Hole-Type IQI Sensitivity Hole-Type Designation 2T Hole DOCUMENTATION The organization shall prepare and document the radiographic technique details. As a minimum, the following information shall be provided. (a) the requirements of Article 1, T-190(a) (b) identification as required by T-224 (c) the dimensional map (if used) of marker placement in accordance with T-275.3 (d) number of exposures (e) X-ray voltage or isotope type used (f) source size (F in T-274.1) (g) base material type and thickness, weld thickness, weld reinforcement thickness, as applicable (h) source-to-object distance (D in T-274.1) (i) distance from source side of object to film (d in T-274.1) (j) film manufacturer and their assigned type/ designation (k) number of film in each film holder/cassette (l) single- or double-wall exposure (m) single- or double-wall viewing If a light image of the “B,” as described in T-223, appears on a darker background of the radiograph, protection from backscatter is insufficient and the radiograph shall be considered unacceptable. A dark image of the “B” on a lighter background is not cause for rejection. T-285 ARTICLE 2 49 ð21Þ ARTICLE 2 ASME BPVC.V-2021 MANDATORY APPENDIX I IN-MOTION RADIOGRAPHY I-210 F = source size: the maximum projected dimension of the radiating source (or focal spot) in the plane perpendicular to the distance b + c from the weld being radiographed w = beam width at the source side of the weld measured in the direction of motion SCOPE In-motion radiography is a technique of film radiography where the object being radiographed and/or the source of radiation is in motion during the exposure. In-motion radiography may be performed on weldments when the following modified provisions to those in Article 2 are satisfied. This Appendix is not applicable to computed radiographic (CR) or digital radiographic (DR) techniques. I-220 GENERAL REQUIREMENTS I-223 BACKSCATTER DETECTION SYMBOL LOCATION NOTE: Use consistent units. CALIBRATION I-263 BEAM WIDTH EXAMINATION I-274 GEOMETRIC AND IN-MOTION UNSHARPNESS I-274.1 Geometric Unsharpness. Geometric unsharpness for in-motion radiography shall be determined in accordance with T-274.1. (a) For longitudinal welds the lead symbol “B” shall be attached to the back of each film cassette or at approximately equal intervals not exceeding 36 in. (914 mm) apart, whichever is smaller. (b) For circumferential welds, the lead symbol “B” shall be attached to the back of the film cassette in each quadrant or spaced no greater than 36 in. (914 mm), whichever is smaller. (c) The lead symbol “B” shall be placed in a location so that it would appear within an area on the radiograph that meets the requirements of T-282. I-260 I-270 I-274.2 In-Motion Unsharpness. In-motion unsharpness of the radiograph shall be determined in accordance with: where D = distance from source of radiation to weld being radiographed d = distance from source side of the weld being radiographed to the film U M = in-motion unsharpness w = beam width at the source side of the weld measured in the direction of motion determined as specified in I-263 The beam width shall be controlled by a metal diaphragm such as lead. The diaphragm for the energy selected shall be at least 10 half value layers thick. The beam width as shown in Figure I-263 shall be determined in accordance with: NOTE: Use consistent units. I-274.3 Unsharpness Limitations. Recommended maximum values for geometric unsharpness and inmotion unsharpness are provided in T-274.2. I-275 LOCATION MARKERS Location markers shall be placed adjacent to the weld at the extremity of each film cassette and also at approximately equal intervals not exceeding 15 in. (381 mm). where a = slit width in diaphragm in direction of motion b = distance from source to the weld side of the diaphragm c = distance from weld side of the diaphragm to the source side of the weld surface I-277 PLACEMENT AND NUMBER OF IQIS (a) For longitudinal welds, hole IQIs shall be placed adjacent to and on each side of the weld seam, or on the weld seam at the beginning and end of the weld seam, 50 ASME BPVC.V-2021 ARTICLE 2 Figure I-263 Beam Width Determination when used, shall be placed on the weld seam so that the length of the wires is across the length of the weld and spaced as indicated above for hole IQIs. and thereafter at approximately equal intervals not exceeding 36 in. (914 mm) or for each film cassette. Wire IQIs, when used, shall be placed on the weld seam so that the length of the wires is across the length of the weld and spaced as indicated above for hole IQIs. (b) For circumferential welds, hole IQIs shall be placed adjacent to and on each side of the weld seam or on the weld seam in each quadrant or spaced no greater than 36 in. (914 mm) apart, whichever is smaller. Wire IQIs, I-279 REPAIRED AREA When radiography of a repaired area is required, the length of the film used shall be at least equal to the length of the original location marker interval. 51 ARTICLE 6 ASME BPVC.V-2021 ARTICLE 6 LIQUID PENETRANT EXAMINATION ð21Þ T-610 SCOPE T-621.3 Minimum and Maximum Step Times. The written procedure shall have minimum and maximum times for the applicable examination steps listed in Table T-621.3. When this Article is specified by a referencing Code Section, the liquid penetrant method described in this Article shall be used together with Article 1, General Requirements. Definitions of terms used in this Article appear in Article 1, Mandatory Appendix I, I-121.3, PT — Liquid Penetrants. T-620 T-630 The term penetrant materials, as used in this Article, is intended to include all penetrants, emulsifiers, solvents or cleaning agents, developers, etc., used in the examination process. The descriptions of the liquid penetrant classifications and material types are provided in SE-165 of Article 24. GENERAL The liquid penetrant examination method is an effective means for detecting discontinuities which are open to the surface of nonporous metals and other materials. Typical discontinuities detectable by this method are cracks, seams, laps, cold shuts, laminations, and porosity. T-640 T-641 In principle, a liquid penetrant is applied to the surface to be examined and allowed to enter discontinuities. All excess penetrant is then removed, the part is dried, and a developer is applied. The developer functions both as a blotter to absorb penetrant that has been trapped in discontinuities, and as a contrasting background to enhance the visibility of penetrant indications. The dyes in penetrants are either color contrast (visible under white light) or fluorescent (visible under ultraviolet light). MISCELLANEOUS REQUIREMENTS CONTROL OF CONTAMINANTS The user of this Article shall obtain certification of contaminant content for all liquid penetrant materials used on nickel base alloys, austenitic or duplex stainless steels, and titanium. These certifications shall include the penetrant manufacturers’ batch numbers and the test results obtained in accordance with Mandatory Appendix II of this Article. These records shall be maintained as required by the referencing Code Section. T-642 T-621 EQUIPMENT WRITTEN PROCEDURE REQUIREMENTS SURFACE PREPARATION (a) In general, satisfactory results may be obtained when the surface of the part is in the as-welded, as-rolled, as-cast, or as-forged condition. Surface preparation by grinding, machining, or other methods may be necessary where surface irregularities could mask indications. (b) Prior to each liquid penetrant examination, the surface to be examined and all adjacent areas within at least 1 in. (25 mm) shall be dry and free of all dirt, grease, lint, scale, welding flux, weld spatter, paint, oil, and other extraneous matter that could obscure surface openings or otherwise interfere with the examination. (c) Typical cleaning agents which may be used are detergents, organic solvents, descaling solutions, and paint removers. Degreasing and ultrasonic cleaning methods may also be used. (d) Cleaning solvents shall meet the requirements of T-641. The cleaning method employed is an important part of the examination process. T-621.1 Requirements. Liquid penetrant examination shall be performed in accordance with a written procedure which shall as a minimum, contain the requirements listed in Table T-621.1. The written procedure shall establish a single value, or range of values, for each requirement. T-621.2 Procedure Qualification. When procedure qualification is specified by the referencing Code Section, a change of a requirement in Table T-621.1 identified as an essential variable shall require requalification of the written procedure by demonstration. A change of a requirement identified as a nonessential variable does not require requalification of the written procedure. All changes of essential or nonessential variables from those specified within the written procedure shall require revision of, or an addendum to, the written procedure. 226 ASME BPVC.V-2021 ARTICLE 6 Table T-621.1 Requirements of a Liquid Penetrant Examination Procedure Requirement Essential Variable Nonessential Variable Identification of and any change in type or family group of penetrant materials including developers, emulsifiers, etc. Surface preparation (finishing and cleaning, including type of cleaning solvent) Method of applying penetrant Method of removing excess surface penetrant Hydrophilic or lipophilic emulsifier concentration and dwell time in dip tanks and agitation time for hydrophilic emulsifiers Hydrophilic emulsifier concentration in spray applications Method of applying developer Minimum and maximum time periods between steps and drying aids Decrease in penetrant dwell time Increase in developer dwell time (Interpretation Time) Minimum light intensity Surface temperature outside 40°F to 125°F (5°C to 52°C) or as previously qualified Performance demonstration, when required Personnel qualification requirements Materials, shapes, or sizes to be examined and the extent of examination Post-examination cleaning technique X … X X X X … … … … X X X X X X X X … … … … … … … … … … … X X X T-650 NOTE: Conditioning of surfaces prior to examination as required in (a) may affect the results. See SE-165, Annex A1. T-651 T-643 DRYING AFTER PREPARATION TECHNIQUE ð21Þ TECHNIQUES Either a color contrast (visible) penetrant or a fluorescent penetrant shall be used with one of the following three penetrant techniques: (a) water washable (b) post-emulsifying (c) solvent removable After cleaning, drying of the surfaces to be examined shall be accomplished by normal evaporation or with forced hot or cold air. A minimum period of time shall be established to ensure that the cleaning solution has evaporated prior to application of the penetrant. Table T-621.3 Minimum and Maximum Time Limits for Steps in Penetrant Examination Procedures Procedure Step Drying after preparation (T-643) Penetrant dwell (T-672) Penetrant removal water washable/solvent removable (T-673.1/T-673.3) Penetrant removal with lipophilic emulsifier [T-673.2(a)] Penetrant removal with hydrophilic emulsifier [T-673.2(b)] Prerinse Immersion Water-emulsifier spray Water immersion or spray post-rinse Drying after penetrant removal (T-674) Solvent removal penetrants Water-washable and post-emulsifiable penetrants Developer application (T-675) Developing and interpretation time (T-675.3 and T-676) 227 Minimum Maximum X X … X … X … X … … … … X X X X … … … X X X X X ARTICLE 6 T-652 ASME BPVC.V-2021 TECHNIQUES FOR STANDARD TEMPERATURES the length of the dwell time, the penetrant shall not be allowed to dry. If for any reason the penetrant does dry, the examination procedure shall be repeated, beginning with a cleaning of the examination surface. As a standard technique, the temperature of the penetrant and the surface of the part to be processed shall not be below 40°F (5°C) nor above 125°F (52°C) throughout the examination period. Local heating or cooling is permitted provided the part temperature remains in the range of 40°F to 125°F (5°C to 52°C) during the examination. Where it is not practical to comply with these temperature limitations, other temperatures and times may be used, provided the procedures are qualified as specified in T-653. T-653 T-673 After the specified penetration (dwell) time has elapsed, any penetrant remaining on the surface shall be removed, taking care to minimize removal of penetrant from discontinuities. T-673.1 Water-Washable Penetrants. (a) Excess water-washable penetrants shall be removed with a water spray. The water pressure shall not exceed 50 psi (350 kPa), and the water temperature shall not exceed 110°F (43°C). (b) As an alternative to (a), water-washable penetrants may be removed by wiping with a clean, dry, lint-free cloth or absorbent paper, repeating the operation until most traces of penetrant have been removed. The remaining traces shall be removed by wiping the surface with a cloth or absorbent paper, lightly moistened with water. To minimize removal of penetrant from discontinuities, care shall be taken to avoid the use of excess water. TECHNIQUES FOR NONSTANDARD TEMPERATURES When it is not practical to conduct a liquid penetrant examination within the temperature range of 40°F to 125°F (5°C to 52°C), the examination procedure at the proposed lower or higher temperature range requires qualification of the penetrant materials and processing in accordance with Mandatory Appendix III of this Article. T-654 TECHNIQUE RESTRICTIONS Fluorescent penetrant examination shall not follow a color contrast penetrant examination. Intermixing of penetrant materials from different families or different manufacturers is not permitted. A retest with water-washable penetrants may cause loss of marginal indications due to contamination. T-660 T-673.2 Post-Emulsification Penetrants. (a) Lipophilic Emulsification. After the required penetrant dwell time, the excess surface penetrant shall be emulsified by immersing or flooding the part with the emulsifier. Emulsification time is dependent on the type of emulsifier and surface condition. The actual emulsification time shall be determined experimentally. After emulsification, the mixture shall be removed by immersing in or rinsing with water. The temperature and pressure of the water shall be as recommended by the manufacturer. (b) Hydrophilic Emulsification. After the required penetrant dwell time, the parts may be prerinsed with water spray or dir ectly immer sed or spr ayed with an emulsifier–water mixture. A prerinse allows removal of excess surface penetrant from examination objects prior to the application of hydrophilic emulsifiers. Hydrophilic emulsifiers work by detergent action. For immersion applications, examination objects must be mechanically moved in the emulsifier bath or the emulsifier must be agitated by air bubbles, so that with either method, the emulsifier comes in contact with the penetrant coating. With immersion, the concentration of the emulsifier– water bath shall be as recommended by the manufacturer. For spray applications, all part surfaces shall be uniformly sprayed with an emulsifier–water mixture. With spray applications, the emulsifier concentration shall be in accordance with the manufacturer’s recommendations, but shall be no greater than 5%. The final step after emulsification is a water immersion or a water spray post-rinse to remove the emulsified penetrant. All dwell times should be kept to a minimum and shall be not more than 2 min unless a longer time is qualified on a specific part. The pressures (water emulsifier and water CALIBRATION Light meters, both visible and fluorescent (black) light meters, shall be calibrated at least once a year or whenever the meter has been repaired. If meters have not been in use for one year or more, calibration shall be done before being used. T-670 T-671 EXAMINATION PENETRANT APPLICATION The penetrant may be applied by any suitable means, such as dipping, brushing, or spraying. If the penetrant is applied by spraying using compressed-air-type apparatus, filters shall be placed on the upstream side near the air inlet to preclude contamination of the penetrant by oil, water, dirt, or sediment that may have collected in the lines. T-672 EXCESS PENETRANT REMOVAL PENETRATION (DWELL) TIME Penetration (dwell) time is critical. The minimum penetration time shall be as required in Table T-672 or as qualified by demonstration for specific applications. The maximum dwell time shall not exceed 2 hr or as qualified by demonstration for specific applications. Regardless of 228 ASME BPVC.V-2021 ARTICLE 6 Table T-672 Minimum Dwell Times Dwell Times [Note (1)], (minutes) Material Aluminum, magnesium, steel, brass and bronze, titanium and high-temperature alloys Carbide-tipped tools Plastic Glass Ceramic Form Type of Discontinuity Castings and welds Cold shuts, porosity, lack of fusion, cracks (all forms) Wrought materials — extrusions, forgings, plate Brazed or welded All forms All forms All forms Laps, cracks Lack of fusion, porosity, cracks Cracks Cracks Cracks Penetrant 5 10 5 5 5 5 NOTE: (1) For temperature range from 50°F to 125°F (10°C to 52°C). For temperatures from 40°F (5°C) up to 50°F (10°C), minimum penetrant dwell time shall be 2 times the value listed. spray) and temperatures (water and emulsifier) shall be in accordance with the requirements for water-washable penetrants. With color contrast penetrants, only a wet developer shall be used. With fluorescent penetrants, a wet or dry developer may be used. NOTE: Additional information may be obtained from SE-165. T-675.1 Dry Developer Application. Dry developer shall be applied only to a dry surface by a soft brush, hand powder bulb, powder gun, or other means, provided the powder is dusted evenly over the entire surface being examined. T-673.3 Solvent Removable Penetrants. Excess solvent removable penetrants shall be removed by wiping with a clean, dry, lint-free cloth or absorbent paper, repeating the operation until most traces of penetrant have been removed. The remaining traces shall be removed by wiping the surface with cloth or absorbent paper, lightly moistened with solvent. To minimize removal of penetrant from discontinuities, care shall be taken to avoid the use of excess solvent. T-675.2 Wet Developer Application. Prior to applying suspension type wet developer to the surface, the developer must be thoroughly agitated to ensure adequate dispersion of suspended particles. (a) Aqueous Developer Application. Aqueous developer may be applied to either a wet or dry surface. It shall be applied by dipping, brushing, spraying, or other means, provided a thin coating is obtained over the entire surface being examined. Drying time may be decreased by using warm air, provided the surface temperature of the part is not raised above 125°F (52°C). Blotting is not permitted. (b) Nonaqueous Developer Application. Nonaqueous developers shall be applied by spraying, except where safety or restricted access preclude it. Under such conditions, developer may be applied by brushing. For waterwashable or post-emulsifiable penetrants, the developer shall be applied to a dry surface. For solvent removable penetrants, the developer may be applied as soon as practical after excess penetrant removal. Drying shall be by normal evaporation. WARNING: Flushing the surface with solvent, following the application of the penetrant and prior to developing, is prohibited. T-674 DRYING AFTER EXCESS PENETRANT REMOVAL (a) For the water-washable or post-emulsifying technique, the surfaces may be dried by blotting with clean materials or by using circulating air, provided the temperature of the surface is not raised above 125°F (52°C). (b) For the solvent removable technique, the surfaces may be dried by normal evaporation, blotting, wiping, or forced air. T-675 DEVELOPING T-675.3 Developing Time. Developing time for final interpretation begins immediately after the application of a dry developer or as soon as a wet developer coating is dry. The developer shall be applied as soon as possible after penetrant removal; the time interval shall not exceed that established in the procedure. Insufficient coating thickness may not draw the penetrant out of discontinuities; conversely, excessive coating thickness may mask indications. 229 ARTICLE 6 T-676 ASME BPVC.V-2021 INTERPRETATION (f) The UV-A light intensity shall be measured with a UV-A light meter prior to use, whenever the light’s power source is interrupted or changed, and at the completion of the examination or series of examinations. (g) Mercury vapor arc lamps produce UV-A wavelengths mainly at a peak wavelength of 365 nm for inducing fluorescence. Light-emitting diode (LED) UV-A sources using a single UV-A LED or an array of UV-A LEDs shall have emission characteristics comparable to those of other UV-A sources. LED UV-A sources shall meet the requirements of SE-2297 and SE-3022. LED UV-A light sources shall be certified as meeting the requirements of SE-3022 and/or ASTM E3022. T-676.1 Final Interpretation. Final interpretation shall be made not less than 10 min nor more than 60 min after the requirements of T-675.3 are satisfied, unless otherwise qualified under T-653. If bleed-out does not alter the examination results, longer periods are permitted. If the surface to be examined is large enough to preclude complete examination within the prescribed or established time, the examination shall be performed in increments. T-676.2 Characterizing Indication(s). The type of discontinuities are difficult to evaluate if the penetrant diffuses excessively into the developer. If this condition occurs, close observation of the formation of indication (s) during application of the developer may assist in characterizing and determining the extent of the indication(s). T-677 POST-EXAMINATION CLEANING When post-examination cleaning is required by the procedure, it should be conducted as soon as practical after Evaluation and Documentation using a process that does not adversely affect the part. T-676.3 Color Contrast Penetrants. With a color contrast penetrant, the developer forms a reasonably uniform white coating. Surface discontinuities are indicated by bleed-out of the penetrant which is normally a deep red color that stains the developer. Indications with a light pink color may indicate excessive cleaning. Inadequate cleaning may leave an excessive background making interpretation difficult. Illumination (natural or supplemental white light) of the examination surface is required for the evaluation of indications. The minimum light intensity shall be 100 fc (1 076 lx). The light intensity, natural or supplemental white light source, shall be measured with a white light meter prior to the evaluation of indications or a verified light source shall be used. Verification of light sources is required to be demonstrated only one time, documented, and maintained on file. T-680 EVALUATION (a) All indications shall be evaluated in terms of the acceptance standards of the referencing Code Section. (b) Discontinuities at the surface will be indicated by bleed-out of penetrant; however, localized surface irregularities due to machining marks or other surface conditions may produce false indications. (c) Broad areas of fluorescence or pigmentation which could mask indications of discontinuities are unacceptable, and such areas shall be cleaned and reexamined. T-676.4 Fluorescent Penetrants. With fluorescent penetrants, the process is essentially the same as in T-676.3, with the exception that the examination is performed using an ultraviolet light, called UV-A light. The examination shall be performed as follows: (a) It shall be performed in a darkened area with a maximum ambient white light level of 2 fc (21.5 lx) measured with a calibrated white light meter at the examination surface. (b) Examiners shall be in a darkened area for at least 5 min prior to performing examinations to enable their eyes to adapt to dark viewing. Glasses or lenses worn by examiners shall not be photosensitive. (c) The examination area shall be illuminated with UV-A lights that operate in the range between 320 nm and 400 nm. (d) U V - A l i g h t s s h a l l a c h i e v e a m i n i m u m o f 1000 μW/cm2 on the surface of the part being examined throughout the examination. (e) Reflectors and filters should be checked and, if necessary, cleaned prior to use. Cracked or broken reflectors, filters, glasses, or lenses shall be replaced immediately. T-690 T-691 DOCUMENTATION RECORDING OF INDICATIONS T-691.1 Nonrejectable Indications. Nonrejectable indications shall be recorded as specified by the referencing Code Section. T-691.2 Rejectable Indications. Rejectable indications shall be recorded. As a minimum, the type of indications (linear or rounded), location and extent (length or diameter or aligned) shall be recorded. T-692 EXAMINATION RECORDS For each examination, the following information shall be recorded: (a) the requirements of Article 1, T-190(a); (b) liquid penetrant type (visible or fluorescent); (c) type (number or letter designation) of each penetrant, penetrant remover, emulsifier, and developer used; (d) map or record of indications per T-691; (e) material and thickness, and; (f) lighting equipment. 230 ASME BPVC.V-2021 ARTICLE 6 MANDATORY APPENDIX II CONTROL OF CONTAMINANTS FOR LIQUID PENETRANT EXAMINATION II-610 SCOPE SE-165, Annex 2 for chlorine and SE-165, Annex 3 for fluorine. The total chlorine and fluorine content shall not exceed 0.1% by weight. This Appendix contains requirements for the control of contaminant content for all liquid penetrant materials used on nickel base alloys, austenitic stainless steels, and titanium. II-640 II-641 II-643 (a) For water used in precleaning or as part of processes that involve water, if potable water (e.g., drinking, bottled, distilled, or deionized water) is used, it is not required to be analyzed for chlorine and sulfur. (b) Any other type of water used that does not meet the requirements of (a) above shall be analyzed for chlorine in accordance with ASTM D1253 and for sulfur in accordance with SD-516. The chlorine content shall not exceed 0.1% by weight and the sulfur content shall not exceed 0.1% by weight. REQUIREMENTS NICKEL BASE ALLOYS When examining nickel base alloys, all penetrant materials shall be analyzed individually for sulfur content in accordance with SE-165, Annex 4. Alternatively, the material may be decomposed in accordance with SD-129 and analyzed in accordance with SD-516. The sulfur content shall not exceed 0.1% by weight. II-642 WATER AUSTENITIC OR DUPLEX STAINLESS STEEL AND TITANIUM II-690 When examining austenitic or duplex stainless steel and titanium, all penetrant materials shall be analyzed individually for chlorine and fluorine content in accordance with SE-165, Annex 4. Alternatively, the material may be decomposed and analyzed in accordance with SD-808 or DOCUMENTATION Certifications obtained on penetrant materials shall include the penetrant manufacturers’ batch numbers and the test results obtained in accordance with II-640. These records shall be maintained as required by the referencing Code Section. 231 ARTICLE 6 ASME BPVC.V-2021 MANDATORY APPENDIX III QUALIFICATION TECHNIQUES FOR EXAMINATIONS AT NONSTANDARD TEMPERATURES SCOPE range. The indications of cracks shall be compared between blocks “A” and “B.” If the indications obtained under the proposed conditions on block “B” are essentially the same as obtained on block “A” during examination at 40°F to 125°F (5°C to 52°C), the proposed procedure shall be considered qualified for use. A procedure qualified at a temperature lower than 40°F (5°C) shall be qualified from that temperature to 40°F (5°C). When a liquid penetrant examination cannot be conducted within the standard temperature range of 40°F to 125°F (5°C to 52°C), the temperature of the examination shall be qualified in accordance with this Appendix. MATERIALS III-641.2 Temperature Greater Than 125°F (52°C). If the proposed temperature for the examination is above 125°F (52°C), block “B” shall be held at this temperature throughout the examination. The indications of cracks shall be compared as described in III-641.1 while block “B” is at the proposed temperature and block “A” is at the 40°F to 125°F (5°C to 52°C) temperature range. A liquid penetrant comparator block shall be made as follows. The liquid penetrant comparator blocks shall be made of aluminum, ASTM B209, Type 2024, 3/8 in. (10 mm) thick, and should have approximate face dimensions of 2 in. × 3 in. (50 mm × 75 mm). At the center of each face, an area approximately 1 in. (25 mm) in diameter shall be marked with a 950°F (510°C) temperatureindicating crayon or paint. The marked area shall be heated with a blowtorch, a Bunsen burner, or similar device to a temperature between 950°F (510°C) and 975°F (524°C). The specimen shall then be immediately quenched in cold water, which produces a network of fine cracks on each face. The block shall then be dried by heating to approximately 300°F (149°C). After cooling, the block shall be cut in half. One-half of the specimen shall be designated block “A” and the other block “B” for identification in subsequent processing. Figure III-630 illustrates the comparator blocks “A” and “B.” As an alternate to cutting the block in half to make blocks “A” and “B,” separate blocks 2 in. × 3 in. (50 mm × 75 mm) can be made using the heating and quenching technique as described above. Two comparator blocks with closely matched crack patterns may be used. The blocks shall be marked “A” and “B.” III-640 III-641 3 in. (75 mm) Figure III-630 Liquid Penetrant Comparator Scribe line 11/2 in. (39 mm) III-630 11/2 in. (39 mm) III-610 2 in. (50 mm) 3/ in. 8 (10 mm) REQUIREMENTS COMPARATOR APPLICATION B III-641.1 Temperature Less Than 40°F (5°C). If it is desired to qualify a liquid penetrant examination procedure at a temperature of less than 40°F (5°C), the proposed procedure shall be applied to block “B” after the block and all materials have been cooled and held at the proposed examination temperature until the comparison is completed. A standard procedure which has previously been demonstrated as suitable for use shall be applied to block “A” in the 40°F to 125°F (5°C to 52°C) temperature A GENERAL NOTE: Dimensions given are for guidance only and are not critical. 232 ASME BPVC.V-2021 ARTICLE 6 indications on the comparator shall be demonstrated at 200°F to 400°F (93°C to 204°C) using the maximum observed dwell time.] To qualify a procedure for temperatures above 125°F (52°C), for penetrants normally used in the 40°F to 125°F (5°C to 52°C) temperature range, the upper temperature limit shall be qualified and the procedure then is usable between the qualified upper temperature and the normal lower temperature of 40°F (5°C). [As an example, to qualify a penetrant normally used in the 40°F to 125°F (5°C to 52°C) temperature range at 200°F (93°C), the capability of the penetrant need only be qualified for 40°F to 200°F (5°C to 93°C) using the normal range dwell times.] The temperature range can be any range desired by the user. For a high-temperature penetrant not normally used in the 40°F to 125°F (5°C to 52°C) temperature range, the capability of a penetrant to reveal indications on the comparator shall be demonstrated at both the lower and upper temperatures. [As an example, to qualify a hightemperature penetrant for use from 200°F to 400°F (93°C to 204°C), the capability of the penetrant to reveal III-641.3 Alternate Techniques for Color Contrast Penetrants. As an alternate to the requirements of III-641.1 and III-641.2, when using color contrast penetrants, it is permissible to use a single comparator block for the standard and nonstandard temperatures and to make the comparison by photography. (a) When the single comparator block and photographic technique is used, the processing details (as applicable) described in III-641.1 and III-641.2 apply. The block shall be thoroughly cleaned between the two processing steps. Photographs shall be taken after processing at the nonstandard temperature and then after processing at the standard temperature. The indication of cracks shall be compared between the two photographs. The same criteria for qualification as III-641.1 shall apply. (b) Identical photographic techniques shall be used to make the comparison photographs. 233 ARTICLE 7 ASME BPVC.V-2021 ARTICLE 7 MAGNETIC PARTICLE EXAMINATION T-710 SCOPE requirements listed in Table T-721. The written procedure shall establish a single value, or range of values, for each requirement. When specified by the referencing Code Section, the magnetic particle examination techniques described in this Article shall be used. In general, this Article is in conformance with SE-709, Standard Guide for Magnetic Particle Testing. This document provides details to be considered in the procedures used. When this Article is specified by a referencing Code Section, the magnetic particle method described in this Article shall be used together with Article 1, General Requirements. Definition of terms used in this Article are in Article 1, Mandatory Appendix I, I-121.4, MT — Magnetic Particle. T-721.2 Procedure Qualification. When procedure qualification is specified by the referencing Code Section, a change of a requirement in Table T-721 identified as an essential variable shall require requalification of the written procedure by demonstration. A change of a requirement identified as a nonessential variable does not require requalification of the written procedure. All changes of essential or nonessential variables from those specified within the written procedure shall require revision of, or an addendum to, the written procedure. T-730 T-720 GENERAL A suitable and appropriate means for producing the necessary magnetic flux in the part shall be employed, using one or more of the techniques listed in and described in T-750. The magnetic particle examination method is applied to detect cracks and other discontinuities on the surfaces of ferromagnetic materials. The sensitivity is greatest for surface discontinuities and diminishes rapidly with increasing depth of discontinuities below the surface. Typical types of discontinuities that can be detected by this method are cracks, laps, seams, cold shuts, and laminations. In principle, this method involves magnetizing an area to be examined, and applying ferromagnetic particles (the examination’s medium) to the surface. Particle patterns form on the surface where the magnetic field is forced out of the part and over discontinuities to cause a leakage field that attracts the particles. Particle patterns are usually characteristic of the type of discontinuity that is detected. Whichever technique is used to produce the magnetic flux in the part, maximum sensitivity will be to linear discontinuities oriented perpendicular to the lines of flux. For optimum effectiveness in detecting all types of discontinuities, each area is to be examined at least twice, with the lines of flux during one examination being approximately perpendicular to the lines of flux during the other. T-721 EQUIPMENT T-731 EXAMINATION MEDIUM The finely divided ferromagnetic particles used for the examination shall meet the following requirements. (a) Particle Types. The particles shall be treated to impart color (fluorescent pigments, nonfluorescent pigments, or both) in order to make them highly visible (contrasting) against the background of the surface being examined. (b) Particles. Dry and wet particles and suspension vehicles shall be in accordance with the applicable specifications listed in SE-709, para. 2.2. (c) Temperature Limitations. Particles shall be used within the temperature range limitations set by the manufacturer of the particles. Alternatively, particles may be used outside the particle manufacturer’s recommendations providing the procedure is qualified in accordance with Article 1, T-150 at the proposed temperature. WRITTEN PROCEDURE REQUIREMENTS T-721.1 Requirements. Magnetic particle examination shall be performed in accordance with a written procedure, which shall, as a minimum, contain the 234 ASME BPVC.V-2021 ARTICLE 7 Table T-721 Requirements of a Magnetic Particle Examination Procedure Requirement Essential Variable Nonessential Variable Magnetizing technique Magnetizing current type or amperage outside range specified by this Article or as previously qualified Surface preparation Magnetic particles (fluorescent/visible, color, particle size, wet/dry) Method of particle application Method of excess particle removal Minimum light intensity Existing coatings, greater than the thickness demonstrated Nonmagnetic surface contrast enhancement, when utilized Performance demonstration, when required Examination part surface temperature outside of the temperature range recommended by the manufacturer of the particles or as previously qualified Shape or size of the examination object Equipment of the same type Temperature (within those specified by manufacturer or as previously qualified) Demagnetizing technique Post-examination cleaning technique Personnel qualification requirements X X … … X X X X X X X X X … … … … … … … … … … … … … … … X X X X X X T-740 T-741 MISCELLANEOUS REQUIREMENTS enhancement. Thickness measurement of this nonmagnetic surface contrast enhancement is not required. SURFACE CONDITIONING T-741.1 NOTE: Refer to T-150(a) for guidance for the demonstration required in T-741.1(d) and T-741.2. Preparation. (a) Satisfactory results are usually obtained when the surfaces are in the as-welded, as-rolled, as-cast, or asforged conditions. However, surface preparation by grinding or machining may be necessary where surface irregularities could mask indications due to discontinuities. T-750 T-751 TECHNIQUE TECHNIQUES One or more of the following five magnetization techniques shall be used: (a) prod technique (b) longitudinal magnetization technique (c) circular magnetization technique (d) yoke technique (e) multidirectional magnetization technique (b) Prior to magnetic particle examination, the surface to be examined and all adjacent areas within at least 1 in. (25 mm) shall be dry and free of all dirt, grease, lint, scale, welding flux and spatter, oil, or other extraneous matter that could interfere with the examination. (c) Cleaning may be accomplished using detergents, organic solvents, descaling solutions, paint removers, vapor degreasing, sand or grit blasting, or ultrasonic cleaning methods. T-752 (d) If nonmagnetic coatings are left on the part in the area being examined, it shall be demonstrated that indications can be detected through the existing maximum coating thickness applied. When AC yoke technique is used, the demonstration shall be in accordance with Mandatory Appendix I of this Article. PROD TECHNIQUE T-752.1 Magnetizing Procedure. For the prod technique, magnetization is accomplished by portable prod type electrical contacts pressed against the surface in the area to be examined. To avoid arcing, a remote control switch, which may be built into the prod handles, shall be provided to permit the current to be applied after the prods have been properly positioned. T-741.2 Nonmagnetic Surface Contrast Enhancement. Nonmagnetic surface contrasts may be applied by the examiner to uncoated surfaces, only in amounts sufficient to enhance particle contrast. When nonmagnetic surface contrast enhancement is used, it shall be demonstrated that indications can be detected through the T-752.2 Magnetizing Current. Direct or rectified magnetizing current shall be used. The current shall be 100 (minimum) amp/in. (4 amp/mm) to 125 (maximum) amp/in. (5 amp/mm) of prod spacing for sections 3/4 in. (19 mm) thick or greater. For sections 235 ARTICLE 7 ASME BPVC.V-2021 less than 3/4 in. (19 mm) thick, the current shall be 90 amp/in. (3.6 amp/mm) to 110 amp/in. (4.4 amp/mm) of prod spacing. (b) Parts With L/D Ratios Less Than 4 but Not Less Than 2. The magnetizing ampere-turns shall be within 10% of the ampere-turns’ value determined as follows: T-752.3 Prod Spacing. Prod spacing shall not exceed 8 in. (200 mm). Shorter spacing may be used to accommodate the geometric limitations of the area being examined or to increase the sensitivity, but prod spacings of less than 3 in. (75 mm) are usually not practical due to banding of the particles around the prods. The prod tips shall be kept clean and dressed. If the open circuit voltage of the magnetizing current source is greater than 25 V, lead, steel, or aluminum (rather than copper) tipped prods are recommended to avoid copper deposits on the part being examined. T-753 (c) Parts With L/D Ratios Less Than 2. Coil magnetization technique cannot be used. (d) If the area to be magnetized extends beyond 9 in. (225 mm) on either side of the coil’s center, field adequacy shall be demonstrated using a magnetic field indicator or artificial flaw shims per T-764. (e) F or la r ge parts du e to s ize and sha pe, the magnetizing current shall be 1200 ampere-turns to 4500 ampere-turns. The field adequacy shall be demonstrated using artificial flaw shims or a pie-shaped magnetic field indicator in accordance with T-764. A Hall-Effect probe gaussmeter shall not be used with encircling coil magnetization techniques. LONGITUDINAL MAGNETIZATION TECHNIQUE T-753.1 Magnetizing Procedure. For this technique, magnetization is accomplished by passing current through a multi-turn fixed coil (or cables) that is wrapped around the part or section of the part to be examined. This produces a longitudinal magnetic field parallel to the axis of the coil. T-753.3 Magnetizing Current. The current required to obtain the necessary magnetizing field strength shall be determined by dividing the ampere-turns obtained in steps T-753.2(a) or T-753.2(b) by the number of turns in the coil as follows: If a fixed, prewound coil is used, the part shall be placed near the side of the coil during inspection. This is of special importance when the coil opening is more than 10 times the cross-sectional area of the part. For example, if a 5-turn coil is used and the ampereturns required are 5000, use T-753.2 Magnetic Field Strength. Direct or rectified current shall be used to magnetize parts examined by this technique. The required field strength shall be calculated based on the length L and the diameter D of the part in accordance with (a) and (b), or as established in (d) and (e), below. Long parts shall be examined in sections not to exceed 18 in. (450 mm), and 18 in. (450 mm) shall be used for the part L in calculating the required field strength. For noncylindrical parts, D shall be the maximum cross-sectional diagonal. T-754 CIRCULAR MAGNETIZATION TECHNIQUE T-754.1 Direct Contact Technique. (a) Magnetizing Procedure. For this technique, magnetization is accomplished by passing current through the part to be examined. This produces a circular magnetic field that is approximately perpendicular to the direction of current flow in the part. (b) Magnetizing Current. Direct or rectified (half-wave rectified or full-wave rectified) magnetizing current shall be used. (1) The current shall be 300 amp/in. (12 A/mm) to 800 amp/in. (31 A/mm) of outer diameter. (2) For parts with geometric shapes other than round, the greatest cross-sectional diagonal in a plane at right angles to the current flow shall be used in lieu of the outer diameter in (1) above. (3) If the current levels required for (1) cannot be obtained, the maximum current obtainable shall be used and the field adequacy shall be demonstrated in accordance with T-764. (a) Parts With L/D Ratios Equal to or Greater Than 4. The magnetizing current shall be within 10% of the ampere-turns’ value determined as follows: For example, a part 10 in. (250 mm) long × 2 in. (50 mm) diameter has an L/D ratio of 5. Therefore, 236 ASME BPVC.V-2021 T-754.2 Central Conductor Technique. ARTICLE 7 Figure T-754.2.2 The Effective Region of Examination When Using an Offset Central Conductor (a) Magnetizing Procedure. For this technique, a central conductor is used to examine the internal surfaces of cylindrically or ring-shaped parts. The central conductor technique may also be used for examining the outside surfaces of these shapes. Where large diameter cylinders are to be examined, the conductor shall be positioned close to the internal surface of the cylinder. When the conductor is not centered, the circumference of the cylinder shall be examined in increments. Field strength measurements in accordance with T-764 shall be used, to determine the extent of the arc that may be examined for each conductor position or the rules in (c) below may be followed. Bars or cables, passed through the bore of a cylinder, may be used to induce circular magnetization. Effective region Central conductor 4d d (b) Magnetizing Current. The field strength required shall be equal to that determined in T-754.1(b) for a single-turn central conductor. The magnetic field will increase in proportion to the number of times the central conductor cable passes through a hollow part. For example, if 6000 A are required to examine a part using a single pass central conductor, then 3000 A are required when 2 passes of the through-cable are used, and 1200 A are required if 5 passes are used (see Figure T-754.2.1). When the central conductor technique is used, magnetic field adequacy shall be verified using a magnetic particle field indicator in accordance with T-764. T-755 YOKE TECHNIQUE For this technique, alternating or direct current electromagnetic yokes, or permanent magnet yokes, shall be used. T-756 MULTIDIRECTIONAL MAGNETIZATION TECHNIQUE T-756.1 Magnetizing Procedure. For this technique, magnetization is accomplished by high amperage power packs operating as many as three circuits that are energized one at a time in rapid succession. The effect of these rapidly alternating magnetizing currents is to produce an overall magnetization of the part in multiple directions. Circular or longitudinal magnetic fields may be generated in any combination using the various techniques described in T-753 and T-754. (c) Offset Central Conductor. When the conductor passing through the inside of the part is placed against an inside wall of the part, the current levels, as given in T-754.1(b)(1) shall apply, except that the diameter used for current calculations shall be the sum of the diameter of the central conductor and twice the wall thickness. The distance along the part circumference (exterior) that is effectively magnetized shall be taken as four times the diameter of the central conductor, as illustrated in Figure T-754.2.2. The entire circumference shall be inspected by rotating the part on the conductor, allowing for approximately a 10% magnetic field overlap. T-756.2 Magnetic Field Strength. Only three phase, full-wave rectified current shall be used to magnetize the part. The initial magnetizing current requirements Figure T-754.2.1 Single-Pass and Two-Pass Central Conductor Technique 237 ARTICLE 7 ASME BPVC.V-2021 T-763 for each circuit shall be established using the previously described guidelines (see T-753 and T-754). The adequacy of the magnetic field shall be demonstrated using artificial flaw shims or a pie-shaped magnetic particle field indicator in accordance with T-764. A Hall-Effect probe gaussmeter shall not be used to measure field adequacy for the multidirectional magnetization technique. An adequate field shall be obtained in at least two nearly perpendicular directions, and the field intensities shall be balanced so that a strong field in one direction does not overwhelm the field in the other direction. For areas where adequate field strengths cannot be demonstrated, additional magnetic particle techniques shall be used to obtain the required two-directional coverage. T-760 T-761 Hall-Effect probe gaussmeters used to verify magnetizing field strength in accordance with T-754 shall be calibrated at least once a year or whenever the equipment has been subjected to a major repair, periodic overhaul, or damage. If equipment has not been in use for a year or more, calibration shall be done prior to first use. T-764 MAGNETIC FIELD ADEQUACY AND DIRECTION T-764.1 Application. The use of magnetic field indicators, artificial shims, or Hall-Effect tangential-field probes are only permitted when specifically referenced by the following magnetizing techniques: (a) Longitudinal (T-753) (b) Circular (T-754) (c) Multidirectional (T-756) CALIBRATION FREQUENCY OF CALIBRATION T-761.1 Magnetizing Equipment. (a) Frequency. Magnetizing equipment with an ammeter shall be calibrated at least once a year, or whenever the equipment has been subjected to major electric repair, periodic overhaul, or damage. If equipment has not been in use for a year or more, calibration shall be done prior to first use. (b) Procedure. The accuracy of the unit’s meter shall be verified annually by equipment traceable to a national standard. Comparative readings shall be taken for at least three different current output levels encompassing the usable range. (c) Tolerance. The unit’s meter reading shall not deviate by more than ±10% of full scale, relative to the actual current value as shown by the test meter. T-764.2 Magnetic Field Adequacy. The applied magnetic field shall have sufficient strength to produce satisfactory indications, but shall not be so strong that it causes masking of relevant indications by nonrelevant accumulations of magnetic particles. Factors that influence the required field strength include the size, shape, and material permeability of the part; the technique of magnetization; coatings; the method of particle application; and the type and location of discontinuities to be detected. When it is necessary to verify the adequacy of magnetic field strength, it shall be verified by using one or more of the following three methods. (a) Pie-Shaped Magnetic Particle Field Indicator. The indicator, shown in Figure T-764.2(a), shall be positioned on the surface to be examined, such that the copperplated side is away from the inspected surface. A suitable field strength is indicated when a clearly defined line (or lines) of magnetic particles form(s) across the copper face of the indicator when the magnetic particles are applied simultaneously with the magnetizing force. When a clearly defined line of particles is not formed, the magnetizing technique shall be changed as needed. Pie-type indicators are best used with dry particle procedures. (b) Artificial Flaw Shims. One of the shims shown in Figure T-764.2(b)(1) or Figure T-764.2(b)(2) whose orientation is such that it can have a component perpendicular to the applied magnetic field shall be used. Shims with linear notches shall be oriented so that at least one notch is perpendicular to the applied magnetic field. Shims with only circular notches may be used in any orientation. Shims shall be attached to the surface to be examined, such that the artificial flaw side of the shim is toward the inspected surface. A suitable field strength is indicated when a clearly defined line (or lines) of magnetic particles, representing the 30% depth flaw, appear (s) on the shim face when magnetic particles are applied simultaneously with the magnetizing force. When a T-761.2 Light Meters. Light meters shall be calibrated at least once a year or whenever a meter has been repaired. If meters have not been in use for one year or more, calibration shall be done before being used. T-762 GAUSSMETERS LIFTING POWER OF YOKES (a) The magnetizing power of yokes shall be verified prior to use each day the yoke is used. The magnetizing power of yokes shall be verified whenever the yoke has been damaged or repaired. (b) Each alternating current electromagnetic yoke shall have a lifting power of at least 10 lb (4.5 kg) at the maximum pole spacing, with contact similar to what will be used during the examination. (c) Each direct current or permanent magnetic yoke shall have a lifting power of at least 40 lb (18 kg) at the maximum pole spacing, with contact similar to what will be used during the examination. (d) Each weight shall be weighed with a scale from a reputable manufacturer and stenciled with the applicable nominal weight prior to first use. A weight need only be verified again if damaged in a manner that could have caused potential loss of material. 238 ASME BPVC.V-2021 T-765 Figure T-764.2(a) Pie-Shaped Magnetic Particle Field Indicator ARTICLE 7 WET PARTICLE CONCENTRATION AND CONTAMINATION Wet Horizontal Units shall have the bath concentration and bath contamination determined by measuring its settling volume. This is accomplished through the use of a pear-shaped centrifuge tube with a 1-mL stem (0.05-mL divisions) for fluorescent particle suspensions or a 1.5-mL stem (0.1-mL divisions) for nonfluorescent suspensions (see SE-709, Appendix X5). Before sampling, the suspension should be run through the recirculating system for at least 30 min to ensure thorough mixing of all particles which could have settled on the sump screen and along the sides or bottom of the tank. T-765.1 Concentration. Take a 100-mL portion of the suspension from the hose or nozzle, demagnetize and allow it to settle for approximately 60 min with petroleum distillate suspensions or 30 min with water-based Figure T-764.2(b)(1) Artificial Flaw Shims 0.002 in. (0.06 mm) A 0.75 in. (20 mm) clearly defined line of particles is not formed, the magnetizing technique shall be changed as needed. Shim-type indicators are best used with wet particle procedures. 0.0006 in. (0.015 mm) A NOTE: The circular shims shown in Figure T-764.2(b)(2) illustration (b) also have flaw depths less and greater than 30%. 0.25 in. (6 mm) (c) Hall-Effect Tangential-Field Probe. A gaussmeter and Hall-Effect tangential-field probe shall be used for measuring the peak value of a tangential field. The probe shall be positioned on the surface to be examined, such that the maximum field strength is determined. A suitable field strength is indicated when the measured field is within the range of 30 G to 60 G (2.4 kAm −1 to 4.8 kAm−1) while the magnetizing force is being applied. See Article 7, Nonmandatory Appendix A. 0.005 in. (0.125 mm) typical Type B Section A–A 0.002 in. (0.05 mm) A 0.75 in. (20 mm) 0.0006 in. (0.015 mm) A 0.5 in. (12.5 mm) T-764.3 Magnetic Field Direction. The direction(s) of magnetization shall be determined by particle indications obtained using an indicator or shims as shown in Figure T-764.2(a), Figure T-764.2(b)(1), or Figure T-764.2(b)(2). When a clearly defined line of particles are not formed (a) in the desired direction, or (b) in at least two nearly perpendicular directions for the multidirectional technique Type C 2 in. (50 mm) Defect Division 0.4 in. (10 mm) 0.2 in. (5 mm) the magnetizing technique shall be changed as needed. Type R Section A–A 0.002 in. (0.05 mm) 0.005 in. (0.125 mm) typical 0.0006 in. (0.015 mm) GENERAL NOTE: Above are examples of artificial flaw shims used in magnetic particle inspection system verification (not drawn to scale). The shims are made of low carbon steel (1005 steel foil). The artificial flaw is etched or machined on one side of the foil to a depth of 30% of the foil thickness. 239 ARTICLE 7 ASME BPVC.V-2021 Figure T-764.2(b)(2) Artificial Flaw Shims 0.75 in. (typ) (19.05 mm) 0.25 in. (6.36 mm) 0.75 in. (typ) (19.05 mm) 0.25 in. (6.36 mm) 0.507 in. diam. O.D. (12.88 mm) 0.507 in. diam. O.D. (12.88 mm) 0.007 in. (typ) (0.18 mm) Notches: Depth: 30% 0.0006 in. (0.015 mm) Shim thickness: 0.002 in. (0.05 mm) 230 Shim Type CX-230 430 Shim Type CX-430 0.007 in. (typ) (0.18 mm) Notches: Depth: 30% 0.0012 in. (0.030 mm) Shim thickness: 0.004 in. (0.10 mm) (a) 0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm) 0.007 in. (type) (0.18 mm) 0.007 in. (type) (0.18 mm) 0.507 in. diam. O.D. (12.88 mm) 0.507 in. diam. O.D. (12.88 mm) 0.383 in. diam. O.D. (9.73 mm) 0.383 in. diam. O.D. (9.73 mm) 0.258 in. diam. O.D. (6.55 mm) 0.258 in. diam. O.D. (6.55 mm) Notch depth: 20% 0.0004 in. (0.010 mm) O.D. 30% 0.0006 in. 4-234 (0.015 mm) center Shim Type 3C4-234 40% 0.0008 in. Shim Thickness 0.004 in. (0.102 mm) (0.020 mm) I.D. Notch depth: 20% 0.0004 in. (0.010 mm) O.D. 30% 0.0006 in. 2-234 (0.015 mm) center Shim Type 3C2-234 40% 0.0008 in. Shim Thickness 0.002 in. (0.05 mm) (0.020 mm) I.D. (b) 0.79 in. (typ) (20.06 mm) 0.235 in. (typ) (5.97 mm) 0.79 in. (typ) (20.06 mm) 0.235 in. (typ) (5.97 mm) 0.395 in. (typ) (10.03 mm) 0.20 in. (typ) (5.08 mm) 0.255 in. diam. O.D. (6.48 mm) 0.255 in. diam. O.D. (6.48 mm) 0.006 in. (typ) (0.152 mm) 0.006 in. (typ) (0.152 mm) 0.395 in. (typ) (10.03 mm) 0.20 in. (typ) (5.08 mm) Notch depth: 30% 0.0006 in. (0.015 mm) Notch depth: 30% 0.0012 in. (0.030 mm) 230 430 Shim Thickness 0.002 in. (0.051 mm) Shim Type CX-230 Shim Thickness 0.004 in. (0.102 mm) Shim Type CX4-430 (c) 240 ASME BPVC.V-2021 (b) Using the Test Ring. The test ring (see Figure T-766.1), is circularly magnetized with full-wave rectified AC passing through a central conductor with a 1 in. to 11/4 in. (25 mm to 32 mm) diameter hole located in the ring center. The conductor should have a length greater than 16 in. (400 mm). The currents used shall be 1400 A, 2500 A, and 3400 A. The minimum number of holes shown shall be three, five, and six, respectively. The ring edge should be examined with either black light or visible light, depending on the type of particles involved. This test shall be run at the three amperages if the unit will be used at these or higher amperages. The amperage values stated shall not be exceeded in the test. If the test does not reveal the required number of holes, the equipment shall be taken out of service and the cause of the loss of sensitivity determined and corrected. This test shall be run at least once per week. suspensions before reading. The volume settling out at the bottom of the tube is indicative of the particle concentration in the bath. T-765.2 Settling Volumes. For fluorescent particles, the required settling volume is from 0.1 mL to 0.4 mL in a 100-mL bath sample and from 1.2 mL to 2.4 mL per 100 mL of vehicle for nonfluorescent particles unless otherwise specified by the particle manufacturer. Concentration checks shall be made at least every eight hours. T-765.3 Contamination. Both fluorescent and nonfluorescent suspensions shall be checked periodically for contaminants such as dirt, scale, oil, lint, loose fluorescent pigment, water (in the case of oil suspensions), and particle agglomerates which can adversely affect the performance of the magnetic particle examination process. The test for contamination shall be performed at least once per week. (a) Carrier Contamination. For fluorescent baths, the liquid directly above the precipitate should be examined with fluorescent excitation light. The liquid will have a little fluorescence. Its color can be compared with a freshly made-up sample using the same materials or with an unused sample from the original bath that was retained for this purpose. If the “used” sample is noticeably more fluorescent than the comparison standard, the bath shall be replaced. (b) Particle Contamination. The graduated portion of the tube shall be examined under fluorescent excitation light if the bath is fluorescent and under visible light (for both fluorescent and nonfluorescent particles) for striations or bands, differences in color or appearance. Bands or striations may indicate contamination. If the total volume of the contaminates, including bands or striations exceeds 30% of the volume magnetic particles, or if the liquid is noticeably fluorescent, the bath shall be replaced. T-770 T-771 EXAMINATION PRELIMINARY EXAMINATION Before the magnetic particle examination is conducted, a check of the examination surface shall be conducted to locate any surface discontinuity openings which may not attract and hold magnetic particles because of their width. T-772 DIRECTION OF MAGNETIZATION At least two separate examinations shall be performed on each area. During the second examination, the lines of magnetic flux shall be approximately perpendicular to those used during the first examination. A different technique for magnetization may be used for the second examination. T-773 T-766 ARTICLE 7 SYSTEM PERFORMANCE OF HORIZONTAL UNITS METHOD OF EXAMINATION The ferromagnetic particles used in an examination medium can be either wet or dry, and may be either fluorescent or nonfluorescent. Examination(s) shall be done by the continuous method. (a) Dry Particles. The magnetizing current shall remain on while the examination medium is being applied and while any excess of the examination medium is removed. (b) Wet Particles. The magnetizing current shall be turned on after the particles have been applied. Flow of particles shall stop with the application of current. Wet particles applied from aerosol spray cans or pump sprayers may be applied before and/or during magnetizing current application. Wet particles may be applied during the application of magnetizing current if they are not applied directly to the examination area and are allowed to flow over the examination area or are applied directly to the examination area with low velocities insufficient to remove accumulated particles. The Ketos (Betz) ring specimen (see Figure T-766.1) shall be used in evaluating and comparing the overall performance and sensitivity of both dry and wet, fluorescent and nonfluorescent magnetic particle techniques using a central conductor magnetization technique. (a) Ketos (Betz) Test Ring Material. The tool steel (Ketos) ring should be machined from AISI 01 material in accordance with Figure T-766.1. Either the machined ring or the steel blank should be annealed at 1,650°F (900°C), cooled 50°F (28°C) per hour to 1,000°F (540°C) and then air cooled to ambient temperature to give comparable results using similar rings that have had the same treatment. Material and heat treatment are important variables. Experience indicates controlling the softness of the ring by hardness (90 HRB to 95 HRB) alone is insufficient. 241 ARTICLE 7 Figure T-766.1 Ketos (Betz) Test Ring 3/ in. (19 mm) 4 Typ. 125 1 2 3 4 5 6 11/4 in. (32 mm) 5 in. (125 mm) 7 12 D ASME BPVC.V-2021 8 9 11 10 242 7/ in. 8 (22 mm) Hole Diameter [Note (1)] “D” [Note (2)] 1 2 3 4 5 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 6 7 0.07 (1.8) 0.14 (3.6) 0.21 (5.3) 0.28 (7.1) 0.35 (9.0) 0.42 (10.8) 0.49 (12.6) 0.56 (14.4) 0.63 (16.2) 0.70 (18.0) 0.77 (19.8) 0.84 (21.6) 0.07 (1.8) 8 0.07 (1.8) 9 0.07 (1.8) 10 0.07 (1.8) 11 0.07 (1.8) 12 0.07 (1.8) GENERAL NOTES: (a) All dimensions are ±0.03 in. (±0.8 mm) or as noted in Notes (1) and (2). (b) In the in-text table, all dimensions are in inches, except for the parenthesized values, which are in millimeters. (c) Material is ANSI 01 tool steel from annealed round stock. (d) The ring may be heat treated as follows: Heat to 1,400°F to 1,500°F (760°C to 790°C). Hold at this temperature for 1 hr. Cool to a minimum rate of 40°F/hr (22°C/h) to below 1,000°F (540°C). Furnace or air cool to room temperature. Finish the ring to RMS 25 and protect from corrosion. NOTES: (1) All hole diameters are ±0.005 in. (±0.1 mm.) Hole numbers 8 through 12 are optional. (2) Tolerance on the D distance is ±0.005 in. (±0.1 mm). ASME BPVC.V-2021 T-774 EXAMINATION COVERAGE T-777.2 Fluorescent Magnetic Particles. With fluorescent magnetic particles, the process is essentially the same as in T-777.1, with the exception that the examination is performed using an ultraviolet light, called UV-A light. The examination shall be performed as follows: (a) It shall be performed in a darkened area with a maximum ambient white light level of 2 fc (21.5 lx) measured with a calibrated white light meter at the examination surface. (b) Examiners shall be in a darkened area for at least 5 min prior to performing examinations to enable their eyes to adapt to dark viewing. Glasses or lenses worn by examiners shall not be photosensitive. (c) The examination area shall be illuminated with UV-A lights that operate in the range between 320 nm and 400 nm. (d) U V - A l i g h t s s h a l l a c h i e v e a m i n i m u m o f 1000 μW/cm2 on the surface of the part being examined throughout the examination. (e) Reflectors, filters, glasses, and lenses should be checked and, if necessary, cleaned prior to use. Cracked or broken reflectors, filters, glasses, or lenses shall be replaced immediately. (f) The UV-A light intensity shall be measured with a UV-A light meter prior to use, whenever the light’s power source is interrupted or changed, and at the completion of the examination or series of examinations. (g) Mercury vapor arc lamps produce UV-A wavelengths mainly at a peak wavelength of 365 nm for inducing fluorescence. LED UV-A sources using a single UV-A LED or an array of UV-A LEDs shall have emission characteristics comparable to those of other UV-A sources. LED UV-A sources shall meet the requirements of SE-2297 and SE-3022. LED UV-A light sources shall be certified as meeting the requirements of SE-3022 and/or ASTM E3022. All examinations shall be conducted with sufficient field overlap to ensure 100% coverage at the required sensitivity (T-764). T-775 RECTIFIED CURRENT (a) Whenever direct current is required rectified current may be used. The rectified current for magnetization shall be either three-phase (full-wave rectified) current, or single phase (half-wave rectified) current. (b) The amperage required with three-phase, full-wave rectified current shall be verified by measuring the average current. (c) The amperage required with single-phase (halfwave rectified) current shall be verified by measuring the average current output during the conducting half cycle only. (d) When measuring half-wave rectified current with a direct current test meter, readings shall be multiplied by two. T-776 EXCESS PARTICLE REMOVAL Accumulations of excess dry particles in examinations shall be removed with a light air stream from a bulb or syringe or other source of low pressure dry air. The examination current or power shall be maintained while removing the excess particles. T-777 ARTICLE 7 INTERPRETATION The interpretation shall identify if an indication as false, nonrelevant, or relevant. False and nonrelevant indications shall be proven as false or nonrelevant. Interpretation shall be carried out to identify the locations of indications and the character of the indication. T-777.3 Fluorescent Magnetic Particles With Other Fluorescent Excitation Wavelengths. Alternatively to the requirements in T-777.2, the examinations may be performed using alternate wavelength light sources which cause fluorescence in specific particle coatings. Any alternate light wavelength light sources and specific particle designations used shall be qualified18 in accordance with Mandatory Appendix IV. The examination shall be performed as follows: (a) It shall be performed in a darkened area. (b) Examiners shall be in a darkened area for at least 5 min prior to performing examinations to enable their eyes to adapt to dark viewing. Glasses or lenses worn by examiners shall not be photochromic or exhibit any fluorescence. (c) If the fluorescence excitation light source emits visible light intensities greater than 2 fc (21.5 lx), the examiner shall wear fluorescence-enhancing filter glasses approved by the light source manufacturer for use with that light source. T-777.1 Visible (Color Contrast) Magnetic Particles. Surface discontinuities are indicated by accumulations of magnetic particles which should contrast with the examination surface. The color of the magnetic particles shall be different than the color of the examination surface. Illumination (natural or supplemental white light) of the examination surface is required for the evaluation of indications. The minimum light intensity shall be 100 fc (1 076 lx). The light intensity, natural or supplemental white light source, shall be measured with a white light meter prior to the evaluation of indications or a verified light source shall be used. Verification of light sources is required to be demonstrated only one time, documented, and maintained on file. 243 ARTICLE 7 ASME BPVC.V-2021 (d) The fluorescence excitation light source shall achieve at least the minimum light intensity on the surface of the part throughout the examination as qualified in the tests of Mandatory Appendix IV. (e) Reflectors, filters, glasses, and lenses should be checked and, if necessary, cleaned prior to use. Cracked or broken reflectors, filters, glasses, or lenses shall be replaced immediately. (f) The fluorescence excitation light intensity shall be measured with a suitable fluorescence excitation light meter prior to use, whenever the light’s power source is interrupted or changed, and at the completion of the examination or series of examinations. T-790 T-778 T-792 T-791 DOCUMENTATION MULTIDIRECTIONAL MAGNETIZATION TECHNIQUE SKETCH A technique sketch shall be prepared for each different geometry examined, showing the part geometry, cable arrangement and connections, magnetizing current for each circuit, and the areas of examination where adequate field strengths are obtained. Parts with repetitive geometries, but different dimensions, may be examined using a single sketch provided that the magnetic field strength is adequate when demonstrated in accordance with T-756.2. DEMAGNETIZATION RECORDING OF INDICATIONS When residual magnetism in the part could interfere with subsequent processing or usage, the part shall be demagnetized any time after completion of the examination. T-792.1 Nonrejectable Indications. Nonrejectable indications shall be recorded as specified by the referencing Code Section. T-779 T-792.2 Rejectable Indications. Rejectable indications shall be recorded. As a minimum, the type of indications (linear or rounded), location and extent (length or diameter or aligned) shall be recorded. POST-EXAMINATION CLEANING When post-examination cleaning is required, it should be conducted as soon as practical using a process that does not adversely affect the part. T-780 T-793 EVALUATION EXAMINATION RECORDS For each examination, the following information shall be recorded: (a) the requirements of Article 1, T-190(a) (b) magnetic particle equipment and type of current (c) magnetic particles (visible or fluorescent, wet or dry) (d) map or record of indications per T-792 (e) material and thickness (f) lighting equipment (a) All indications shall be evaluated in terms of the acceptance standards of the referencing Code Section. (b) Discontinuities on or near the surface are indicated by retention of the examination medium. However, localized surface irregularities due to machining marks or other surface conditions may produce false indications. (c) Broad areas of particle accumulation, which might mask indications from discontinuities, are prohibited, and such areas shall be cleaned and reexamined. 244 ASME BPVC.V-2021 ARTICLE 7 MANDATORY APPENDIX I MAGNETIC PARTICLE EXAMINATION USING THE AC YOKE TECHNIQUE ON FERROMAGNETIC MATERIALS COATED WITH NONFERROMAGNETIC COATINGS I-710 SCOPE I-721.2 Procedure Qualification/Technique Validation. When procedure qualification is specified, a change of a requirement in Table T-721 or Table I-721 identified as an essential variable from the specified value, or range of values, shall require requalification of the written procedure and validation of the technique. A change of a requirement identified as an nonessential variable from the specified value, or range of values, does not require requalification of the written procedure. All changes of essential or nonessential variables from the value, or range of values, specified by the written procedure shall require revision of, or an addendum to, the written procedure. This Appendix provides the Magnetic Particle examination methodology and equipment requirements applicable for performing Magnetic Particle examination on ferromagnetic materials with nonferromagnetic coatings. I-720 GENERAL Requirements of Article 7 apply unless modified by this Appendix. I-722 I-721 WRITTEN PROCEDURE REQUIREMENTS PERSONNEL QUALIFICATION Personnel qualification requirements shall be in accordance with the referencing Code Section. I-721.1 Requirements. Magnetic Particle examination shall be performed in accordance with a written procedure which shall, as a minimum, contain the requirements listed in Tables T-721 and I-721. The written procedure shall establish a single value, or range of values, for each requirement. Table I-721 Requirements of AC Yoke Technique on Coated Ferritic Component Requirement Identification of surface configurations to be examined, including coating materials, maximum qualified coating thickness, and product forms (e.g., base material or welded surface) Surface condition requirements and preparation methods Manufacturer and model of AC yoke Manufacturer and type of magnetic particles Minimum and maximum pole separation Identification of the steps in performing the examination Minimum lighting intensity and AC yoke lifting power requirements [as measured in accordance with Technique Qualification (I-721.2)] Methods of identifying flaw indications and discriminating between flaw indications and false or nonrelevant indications (e.g., magnetic writing or particles held by surface irregularities) Instructions for identification and confirmation of suspected flaw indications Applicator other than powder blower Method of measuring coating thickness Recording criteria Personnel qualification requirements unique to this technique Reference to the procedure qualification records 245 Essential Variable Nonessential Variable X … X X X X X X … … … … … … X … X X … … … … … … X X X X ARTICLE 7 I-723 ASME BPVC.V-2021 PROCEDURE/TECHNIQUE DEMONSTRATION shall be the same specification and heat treatment as the coated ferromagnetic material to be examined. As an alternative to the material requirement, other materials and heat treatments may be qualified provided: (1) The measured yoke maximum lifting force on the material to be examined is equal to or greater than the maximum lifting force on the qualification specimen material. Both values shall be determined with the same or comparable equipment and shall be documented as required in (c). (2) All the requirements of (b) through (g) are met for the alternate material. (b) Examine the uncoated specimen in the most unfavorable orientation expected during the performance of the production examination. (c) Document the measured yoke maximum lifting power, illumination levels, and the results. (d) Measure the maximum coating thickness on the item to be examined in accordance with the requirements of I-741. (e) Coat the specimen with the same type of coating, conductive or nonconductive, to the maximum thickness measured on the production item to be examined. Alternately, nonconductive shim stock may be used to simulate nonconductive coatings. (f) Examine the coated specimen in the most unfavorable orientation expected during the performance of the production examination. Document the measured yoke maximum lifting power, illumination level, and examination results. (g) Compare the length of the indication resulting from the longest flaw no longer than the maximum flaw size allowed by the applicable acceptance criteria, before and after coating. The coating thickness is qualified when the length of the indication on the coated surface is at least 50% of the length of the corresponding indication prior to coating. (h) Requalification of the procedure is required for a decrease in either the AC yoke lifting power or the illumination level, or for an increase in the coating thickness. The procedure/technique shall be demonstrated to the satisfaction of the Inspector in accordance with the requirements of the referencing Code Section. I-730 EQUIPMENT (a) The magnetizing equipment shall be in accordance with Article 7. (b) When the dry powder technique is used, a compressed air powder blower shall be utilized for powder application in any position. Other applicators may be used if qualified in the same surface position as the examination object surface. Applicators qualified for the overhead position may be used in any other position. Applicators qualified for the vertical position may be used in the horizontal and flat positions. (c) Magnetic particles shall contrast with the component background. (d) Nonconductive materials such as plastic shim stock may be used to simulate nonconductive nonferromagnetic coatings for procedure and personnel qualification. ð21Þ I-740 MISCELLANEOUS REQUIREMENTS I-741 COATING THICKNESS MEASUREMENT The procedure demonstration and performance of examinations shall be preceded by measurement of the coating thickness in the areas to be examined. If the coating is nonconductive, an eddy current technique or magnetic technique may be used to measure the coating thickness. The magnetic technique shall be in accordance with SD-7091, Standard Practice for Nondestructive Measurement of Dry Film Thickness of Nonmagnetic Coatings Applied to Ferrous Metals and Nonmagnetic, Nonconductive Coatings Applied to Non-Ferrous Metals. When coatings are conductive and nonferromagnetic, a coating thickness technique shall be used in accordance with SD-7091. Coating measurement equipment shall be used in accordance with the equipment manufacturer’s instructions. Coating thickness measurements shall be taken at the intersections of a 2 in. (50 mm) maximum grid pattern over the area of examination and at least onehalf the maximum yoke leg separation beyond the examination area. The thickness shall be the mean of three separate readings within 1/4 in. (6 mm) of each intersection. I-750 I-751 I-760 I-761 CALIBRATION YOKE MAXIMUM LIFTING FORCE The maximum lifting force of the AC yoke shall be determined at the actual leg separation to be used in the examination. This may be accomplished by holding the yoke with a 10 lb (4.5 kg) ferromagnetic weight between the legs of the yoke and adding additional weights, calibrated on a postage or other scale, until the ferromagnetic weight is released. The lifting power of the yoke shall be the combined weight of the ferromagnetic material and the added weights, before the ferromagnetic weight was released. Other methods may be used such as a load cell. TECHNIQUE TECHNIQUE QUALIFICATION (a) A qualification specimen is required. The specimen shall be of similar geometry or weld profile and contain at least one linear surface indication no longer than 1/16 in. (1.5 mm) in length. The material used for the specimen 246 ASME BPVC.V-2021 I-762 LIGHT INTENSITY MEASUREMENT (b) Examine the coated item in accordance with the qualified procedure. The black light or white light intensity (as appropriate) on the surface of the component shall be no less than that used in the qualification test. An appropriate calibrated black light and/or white light meter shall be used for the tests. Minimum white light or black light intensities shall meet the requirements of T-777.1 or T-777.2 as applicable. I-780 EVALUATION If an indication greater than 50% of the maximum allowable flaw size is detected, the coating in the area of the indication shall be removed and the examination repeated. I-762.1 White Light. The white light intensity shall be measured at the inspection surface. The white light intensity for the examination shall be no less than what was used in the qualification. I-762.2 Black Light. The black light intensity shall be measured at the distance from the black light in the procedure qualification and at the same distance on the examination specimen. The black light intensity shall be no less than that used to qualify the procedure. In addition, the maximum white light intensity shall be measured as background light on the inspection surface. The background white light for the examination shall be no greater than what was used in the qualification. I-770 ARTICLE 7 I-790 I-791 DOCUMENTATION EXAMINATION RECORD For each examination, the information required in the records section of T-793 and the following information shall be recorded: (a) identification of the procedure/technique (b) description and drawings or sketches of the qualification specimen, including coating thickness measurements and flaw dimensions EXAMINATION (c) equipment and materials used (d) illumination level and yoke lifting power (a) Surfaces to be examined, and all adjacent areas within at least 1 in. (25 mm), shall be free of all dirt, grease, lint, scale, welding flux and spatter, oil, and loose, blistered, flaking, or peeling coating. (e) qualification results, including maximum coating thickness and flaws detected 247 ASME BPVC.V-2021 ARTICLE 9 ARTICLE 9 VISUAL EXAMINATION T-910 T-922 SCOPE The user of this Article shall be responsible for assigning qualified personnel to perform visual examinations to the requirements of this Article. At the option of the organization, he may maintain one certification for each product, or several separate signed records based on the area or type of work, or both combined. Where impractical to use specialized visual examination personnel, knowledgeable and trained personnel, having limited qualifications, may be used to perform specific examinations, and to sign the report forms. Personnel performing examinations shall be qualified in accordance with requirements of the referencing Code Section. (a) This Article contains methods and requirements for visual examination applicable when specified by a referencing Code Section. Specific visual examination procedures required for every type of examination are not included in this Article, because there are many applications where visual examinations are required. Some examples of these applications include nondestructive examinations, leak testing, in-service examinations and fabrication procedures. (b) The requirements of Article 1, General Requirements, apply when visual examination, in accordance with Article 9, is required by a referencing Code Section. (c) Definitions of terms for visual examination appear in Article 1, Mandatory Appendix I, I-121.6, VT — Visual Examination. T-920 T-921 PERSONNEL REQUIREMENTS T-923 PHYSICAL REQUIREMENTS Personnel shall have an annual vision test to assure natural or corrected near distance acuity such that they are capable of reading standard J-1 letters on standard Jaeger test type charts for near vision. Equivalent near vision tests are acceptable. GENERAL WRITTEN PROCEDURE REQUIREMENTS T-921.1 Requirements. Visual examinations shall be performed in accordance with a written procedure, which shall, as a minimum, contain the requirements listed in Table T-921. The written procedure shall establish a single value, or range of values, for each requirement. Table T-921 Requirements of a Visual Examination Procedure T-921.2 Procedure Qualification. When procedure qualification is specified by the referencing Code Section, a change of a requirement in Table T-921 identified as an essential variable shall require requalification of the written procedure by demonstration. A change of a requirement identified as a nonessential variable does not require requalification of the written procedure. All changes of essential or nonessential variables from those specified within the written procedure shall require revision of, or an addendum to, the written procedure. Requirement (as Applicable) Change in technique used Direct to or from translucent Direct to remote Remote visual aids Personnel performance requirements, when required Lighting intensity (decrease only) Configurations to be examined and base material product forms (pipe, plate, forgings, etc.) Lighting equipment Methods or tools used for surface preparation Equipment or devices used for a direct technique Sequence of examination Personnel qualifications T-921.3 Demonstration. The procedure shall contain or reference a report of what was used to demonstrate that the examination procedure was adequate. In general, a fine line 1/32 in. (0.8 mm) or less in width, an artificial imperfection or a simulated condition, located on the surface or a similar surface to that to be examined, may be considered as a method for procedure demonstration. The condition or artificial imperfection should be in the least discernable location on the area surface to be examined to validate the procedure. 293 Essential Variable Nonessential Variable … X X X … … … … … X X … … X … … X X … X … … X X ARTICLE 9 T-930 ASME BPVC.V-2021 EQUIPMENT Equipment used for visual examination techniques, for example, direct, remote, or translucent, shall have the capabilities as specified in the procedure. Capabilities include, but are not limited to viewing, magnifying, identifying, measuring, and/or recording observations in accordance with requirements of the referencing Code Section. lighting. The illuminator shall provide light of an intensity that will illuminate and diffuse the light evenly through the area or region under examination. The ambient lighting must be so arranged that there are no surface glares or reflections from the surface under examination and shall be less than the light applied through the area or region under examination. The artificial light source shall have sufficient intensity to permit “candling” any translucent laminate thickness variations. T-950 TECHNIQUE T-955 T-951 APPLICATIONS Light meters shall be calibrated at least once a year or whenever they have been repaired. If meters have not been in use for 1 yr or more, they shall be calibrated before they are used. Visual examination is generally used to determine such things as the surface condition of the part, alignment of mating surfaces, shape, or evidence of leaking. In addition, visual examination is used to determine a composite material’s (translucent laminate) subsurface conditions. T-952 T-980 DIRECT VISUAL EXAMINATION T-990 T-991 DOCUMENTATION REPORT OF EXAMINATION (a) A written report of the examination shall contain the following information: (1) the date of the examination (2) procedure identification and revision used (3) technique used (4) results of the examination (5) examination personnel identity, and, when required by the referencing Code Section, qualification level (6) identification of the part or component examined (b) Even though dimensions, etc., were recorded in the process of visual examination to aid in the evaluation, there need not be documentation of each viewing or each dimensional check. Documentation shall include all observation and dimensional checks specified by the referencing Code Section. REMOTE VISUAL EXAMINATION In some cases, remote visual examination may have to be substituted for direct examination. Remote visual examination may use visual aids such as mirrors, telescopes, borescopes, fiber optics, cameras, or other suitable instruments. Such systems shall have a resolution capability and light intensity at least equivalent to that obtainable by direct visual observation. T-954 EVALUATION (a) All examinations shall be evaluated in terms of the acceptance standards of the referencing Code Section. (b) An examination checklist shall be used to plan visual examination and to verify that the required visual observations were performed. This checklist establishes minimum examination requirements and does not indicate the maximum examination which the Manufacturer may perform in process. Direct visual examination may usually be made when access is sufficient to place the eye within 24 in. (600 mm) of the surface to be examined and at an angle not less than 30 deg to the surface to be examined. Mirrors may be used to improve the angle of vision, and aids such as a magnifying lens may be used to assist examinations. Illumination (natural or supplemental white light) of the examination surface is required for the specific part, component, vessel, or section thereof being examined. The minimum light intensity shall be 100 fc (1 076 lx). The light intensity, natural or supplemental white light source, shall be measured with a white light meter prior to the examination or a verified light source shall be used. Verification of light sources is required to be demonstrated only one time, documented, and maintained on file. T-953 LIGHT METER CALIBRATION TRANSLUCENT VISUAL EXAMINATION Translucent visual examination is a supplement of direct visual examination. The method of translucent visual examination uses the aid of artificial lighting, which can be contained in an illuminator that produces directional T-993 RECORD MAINTENANCE Records shall be maintained as required by the referencing Code Section. 294 ASME BPVC.V-2021 ARTICLE 10 ARTICLE 10 LEAK TESTING T-1010 SCOPE T-1020 T-1021 This Article describes methods and requirements for the performance of leak testing. (a) When a leak testing method or technique of Article 10 is specified by a referencing Code Section, the leak test method or technique shall be used together with Article 1, General Requirements. (b) Definition of terms used in this Article are in Article 1, Mandatory Appendix I, I-121.7, LT — Leak Testing. (c) The test methods or techniques of these methods can be used for the location of leaks or the measurement of leakage rates. The specific test method(s) or technique(s) and Glossary of Terms of the methods in this Article are described in Mandatory Appendices I through X of Article 10 as follows: Mandatory Appendix I — Bubble Test — Direct Pressure Technique Mandatory Appendix II — Bubble Test — Vacuum Box Technique Mandatory Appendix III — Halogen Diode Detector Probe Test Mandatory Appendix IV — Helium Mass Spectrometer Test — Detector Probe Technique Mandatory Appendix V — Helium Mass Spectrometer Test — Tracer Probe Technique Mandatory Appendix VI — Pressure Change Test Mandatory Appendix VIII — Thermal Conductivity Detector Probe Test Mandatory Appendix IX — Helium Mass Spectrometer Test — Hood Technique Mandatory Appendix X — Ultrasonic Leak Detector Test Mandatory Appendix XI — Helium Mass Spectrometer — Helium-Filled-Container Leakage Rate Test Nonmandatory Appendix A — Supplementary Leak Testing Equation Symbols GENERAL WRITTEN PROCEDURE REQUIREMENTS T-1021.1 Requirements. Leak testing shall be performed in accordance with a written procedure, which shall, as a minimum, contain the requirements listed in the applicable Appendices, paras. I-1021 through X-1021 and Tables I-1021 through X-1021. The written procedure shall establish a single value, or range of values, for each requirement. T-1021.2 Modification of Requirements. Article 10 contains test techniques; therefore, there are requirements that cannot be modified by the organization through the demonstration process per T-150. Only those requirements listed in Tables I-1021 through X-1021 may be so modified by demonstration. T-1021.3 Procedure Qualification. When procedure qualification is specified by the referencing Code Section, a change of a requirement in the applicable Appendix Tables I-1021 through X-1021 identified as an essential variable shall require requalification of the written procedure by demonstration. A change of a requirement identified as a nonessential variable does not require requalification of the written procedure. All changes of essential and nonessential elements from those specified within the written procedure shall require revision of, or an addendum to, the written procedure. T-1022 REFERENCING CODE For the leak testing method(s) or technique(s) specified by the referencing Code, the referencing Code Section shall then be consulted for the following: (a) personnel qualification/certification (b) technique(s)/calibration standards (c) extent of examination (d) acceptable test sensitivity or leakage rate (e) report requirements (f) retention of records 295 ARTICLE 10 T-1030 T-1031 ASME BPVC.V-2021 EQUIPMENT T-1044 Unless specified in the applicable Mandatory Appendix of this Article or by the referencing Code Section, components that are to be pressure-leak tested shall not be tested at a pressure exceeding 25% of the Design Pressure. GAGES (a) Gage Range. When dial indicating and recording pressure gage(s) are used in leak testing, they should preferably have the dial(s) graduated over a range of approximately double the intended maximum pressure, but in no case shall the range be less than 11/2 nor more than four times that pressure. These range limits do not apply to dial indicating and recording vacuum gages. Range requirements for other types of gages given in an applicable Mandatory Appendix shall be as required by that Appendix. (b) Gage Location. When components are to be pressure/vacuum leak tested, the dial indicating gage(s) shall be connected to the component or to the component from a remote location, with the gage(s) readily visible to the operator controlling the pressure/vacuum throughout the duration of pressurizing, evacuating, testing, and depressurizing or venting of the component. For large vessels or systems where one or more gages are specified or required, a recording type gage is recommended, and it may be substituted for one of the two or more indicating type gages. (c) When other types of gage(s) are required by an applicable Mandatory Appendix, they may be used in conjunction with or in place of dial indicating or recording type gages. T-1040 T-1041 T-1050 T-1051 T-1052 PRELIMINARY LEAK TEST TEST SEQUENCE It is recommended that leak testing be performed before hydrostatic or hydropneumatic testing. T-1060 T-1061 CALIBRATION PRESSURE/VACUUM GAGES (a) All dial indicating and recording type gages used shall be calibrated against a standard deadweight tester, a calibrated master gage, or a mercury column, and recalibrated at least once a year, when in use, unless specified differently by the referencing Code Section or Mandatory Appendix. All gages used shall provide results accurate to within the Manufacturer’s listed accuracy and shall be recalibrated at any time that there is reason to believe they are in error. (b) When other than dial indicating or recording type gages are required by an applicable Mandatory Appendix, they shall be calibrated as required by that Mandatory Appendix or referencing Code Section. MISCELLANEOUS REQUIREMENTS CLEANLINESS T-1062 TEMPERATURE MEASURING DEVICES When temperature measurement is required by the referencing Code Section or Mandatory Appendix, the device(s) shall be calibrated in accordance with the requirements of that Code Section or Mandatory Appendix. OPENINGS All openings shall be sealed using plugs, covers, sealing wax, cement, or other suitable material that can be readily and completely removed after completion of the test. Sealing materials shall be tracer gas free. T-1063 T-1043 PROCEDURE Prior to employing a sensitive leak testing method, it may be expedient to perform a preliminary test to detect and eliminate gross leaks. This shall be done in a manner that will not seal or mask leaks during the specified test. The surface areas to be tested shall be free of oil, grease, paint, or other contaminants that might mask a leak. If liquids are used to clean the component or if a hydrostatic or hydropneumatic test is performed before leak testing, the component shall be dry before leak testing. T-1042 PRESSURE/VACUUM (PRESSURE LIMITS) TEMPERATURE CALIBRATION LEAK STANDARDS T-1063.1 Reservoir Leak Standard. This standard leak shall have a reservoir of the tracer gas connected to the leak. The leak standard shall (a) have a leakage rate in the range and tracer gas species specified by the referencing Code Section or, if not specified, per the Mandatory Appendix. (b) be calibrated with discharge either to vacuum or to an air environment of 1 atm (101 kPa absolute) to match the test application or instrument type. The minimum metal temperature for all components during a test shall be as specified in the applicable Mandatory Appendix of this Article or in the referencing Code Section for the hydrostatic, hydropneumatic, or pneumatic test of the pressure component or parts. The minimum or maximum temperature during the test shall not exceed that temperature compatible with the leak testing method or technique used. 296 ASME BPVC.V-2021 T-1090 T-1063.2 Nonreservoir Leak Standard. This standard leak does not have an inherent supply of tracer gas. The leak shall (a) have a leakage rate in the range and tracer gas species specified by the referencing Code Section or, if not specified, per the Mandatory Appendix. (b) be calibrated with discharge either to vacuum or to an air environment of 1 atm (101 kPa absolute) to match the test application. (c) be calibrated at a pressure differential across the leak of 1 atm (14.7 psi, 101 kPa) or at a differential that represents the differential to be used in the specific test procedure. T-1070 T-1091 T-1081 DOCUMENTATION TEST REPORT The test report shall contain, as a minimum, the following information as applicable to the method or technique: (a) date of test (b) certified level and name of operator (c) test procedure (number) and revision number (d) test method or technique (e) test results (f) component identification (g) test instrument, standard leak, and material identification (h) test conditions, test pressure, tracer gas, and gas concentration (i) gage(s) — manufacturer, model, range, and identification number (j) temperature measuring device(s) and identification number(s) (k) sketch showing method or technique setup TEST See applicable Mandatory Appendix of this Article. T-1080 ARTICLE 10 EVALUATION ACCEPTANCE STANDARDS Unless otherwise specified in the referencing Code Section, the acceptance criteria given for each method or technique of that method shall apply. The supplemental leak testing equations for calculating leakage rates for the method or technique used are stated in the Mandatory Appendices of this Article. T-1092 RECORD RETENTION The test report shall be maintained in accordance with the requirements of the referencing Code Section. 297 ASME BPVC.V-2021 ARTICLE 23, SE-797/SE-797M STANDARD PRACTICE FOR MEASURING THICKNESS BY MANUAL ULTRASONIC PULSE-ECHO CONTACT METHOD SE-797/SE-797M (Identical with ASTM Specification E797/E797M-15.) This document, in whole or in part, is mandatory only to the extent specified in the referencing Article(s) of Subsection A, or as indicated in other Code Sections or referencing documents. 653 ARTICLE 23, SE-797/SE-797M ASME BPVC.V-2021 Standard Practice for Measuring Thickness by Manual Ultrasonic Pulse-Echo Contact Method 1. Scope 2.2 ASNT Documents: Nondestructive Testing Handbook, 2nd Edition, Vol 7 SNT-TC-1A Recommended Practice for Personnel Qualification and Certification in Nondestructive Testing ANSI/ASNT CP-189 Standard for Qualification and Certification of Nondestructive Testing Personnel 2.3 Aerospace Industries Association Document: NAS-410 Certification and Qualification of Nondestructive Testing Personnel 2.4 ISO Standard: ISO 9712 Non-Destructive Testing—Qualification and Certification of NDT Personnel 1.1 This practice provides guidelines for measuring the thickness of materials using the contact pulse-echo method at temperatures not to exceed 93°C [200°F]. 1.2 This practice is applicable to any material in which ultrasonic waves will propagate at a constant velocity throughout the part, and from which back reflections can be obtained and resolved. 1.3 Units—The values stated in either SI units or inchpound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. 3. Terminology 3.1 Definitions: Definitions—For definitions of terms used in this practice, refer to Terminology E1316. 4. Summary of Practice 4.1 Thickness (T), when measured by the pulse-echo ultrasonic method, is a product of the velocity of sound in the material and one half the transit time (round trip) through the material. 2. Referenced Documents 2.1 ASTM Standards: E317 Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use of Electronic Measurement Instruments E494 Practice for Measuring Ultrasonic Velocity in Materials E543 Specification for Agencies Performing Nondestructive Testing E1316 Terminology for Nondestructive Examinations T5 Vt 2 where: T = thickness, V = velocity, and t = transit time. 4.2 The pulse-echo ultrasonic instrument measures the transit time of the ultrasonic pulse through the part. 4.3 The velocity in the material being examined is a function of the physical properties of the material. It is usually assumed to be a constant for a given class of materials. Its approximate value can be obtained from Table X3.1 in Practice 654 ASME BPVC.V-2021 ARTICLE 23, SE-797/SE-797M NOTE 1—Slope of velocity conversion line is approximately that of steel. FIG. 1 Transit Time/Thickness Relationship 5. Significance and Use E494 or from the Nondestructive Testing Handbook, or it can be determined empirically. 5.1 The techniques described provide indirect measurement of thickness of sections of materials not exceeding temperatures of 93°C [200°F]. Measurements are made from one side of the object, without requiring access to the rear surface. 4.4 One or more reference blocks are required having known velocity, or of the same material to be examined, and having thicknesses accurately measured and in the range of thicknesses to be measured. It is generally desirable that the thicknesses be “round numbers” rather than miscellaneous odd values. One block should have a thickness value near the maximum of the range of interest and another block near the minimum thickness. 5.2 Ultrasonic thickness measurements are used extensively on basic shapes and products of many materials, on precision machined parts, and to determine wall thinning in process equipment caused by corrosion and erosion. 5.3 Recommendations for determining the capabilities and limitations of ultrasonic thickness gages for specific applications can be found in the cited references.1,2 4.5 The display element (A-scan display, meter, or digital display) of the instrument must be adjusted to present convenient values of thickness dependent on the range being used. The control for this function may have different names on different instruments, including range, sweep, material standardize, or velocity. 6. Basis of Application 6.1 The following items are subject to contractual agreement between the parties using or referencing this practice. 6.2 Personnel Qualification: 6.2.1 If specified in the contractual agreement, personnel performing examinations to this standard shall be qualified in accordance with a nationally or internationally recognized 4.6 The timing circuits in different instruments use various conversion schemes. A common method is the so-called time/analog conversion in which the time measured by the instrument is converted into a proportional d-c voltage which is then applied to the readout device. Another technique uses a very high-frequency oscillator that is modulated or gated by the appropriate echo indications, the output being used either directly to suitable digital readouts or converted to a voltage for other presentation. A relationship of transit time versus thickness is shown graphically in Fig. 1. 1 Bosselaar, H., and Goosens, J.C.J., “Method to Evaluate Direct-Reading Ultrasonic Pulse-Echo Thickness Meters,” Materials Evaluation, March 1971, pp. 45–50. 2 Fowler, K.A., Elfbaum, G.M., Husarek, V., and Castel, J., “Applications of Precision Ultrasonic Thickness Gaging,” Proceedings of the Eighth World Conference on Nondestructive Testing, Cannes, France, Sept. 6–11, 1976, Paper 3F.5. 655 ARTICLE 23, SE-797/SE-797M ASME BPVC.V-2021 tions. (See Fig. 2.) For optimum performance, it is often necessary that the instrument and search units be matched. NDT personnel qualification practice or standard such as ANSI/ASNT CP-189, SNT-TC-1A, NAS-410, ISO 9712, or a similar document and certified by the employer or certifying agency, as applicable. The practice or standard used and its applicable revision shall be identified in the contractual agreement between the using parties. 7.3 Standardization Blocks—The general requirements for appropriate standardization blocks are given in 4.4, 8.1.3, 8.2.2.1, 8.3.2, and 8.4.3. Multi-step blocks that may be useful for these standardization procedures are described in Appendix X1 (Figs. X1.1 and X1.2). 6.3 Qualification of Nondestructive Agencies—If specified in the contractual agreement, NDT agencies shall be qualified and evaluated as described in Specification E543. The applicable edition of Specification E543 shall be specified in the contractual agreement. 8. Standardization of Apparatus 8.1 Case I—Direct Contact, Single-Element Search Unit: 8.1.1 Conditions—The display start is synchronized to the initial pulse. All display elements are linear. Full thickness is displayed on the A-scan display. 8.1.2 Under these conditions, we can assume that the velocity conversion line effectively pivots about the origin (Fig. 1). It may be necessary to subtract the wear-plate time, requiring minor use of delay control. It is recommended that standardization blocks providing a minimum of two thicknesses that span the thickness range be used to check the full-range accuracy. 8.1.3 Place the search unit on a standardization block of known thickness with suitable couplant and adjust the instrument controls (material standardization, range, sweep, or velocity) until the display presents the appropriate thickness reading. 8.1.4 The readings should then be checked and adjusted on standardization blocks with thickness of lesser value to improve the overall accuracy of the system. 6.4 Procedures and Techniques—The procedures and techniques to be utilized shall be as specified in the contractual agreement. 6.5 Surface Preparation—The pre-examination surface preparation criteria shall be specified in the contractual agreement. 7. Apparatus 7.1 Instruments—Thickness-measurement instruments are divided into three groups: (1) Flaw detectors with an A-scan display readout, (2) Flaw detectors with an A-scan display and direct thickness readout, and (3) Direct thickness readout. 7.1.1 Flaw detectors with A-scan display readouts display time/amplitude information. Thickness determinations are made by reading the distance between the zero-corrected initial pulse and first-returned echo (back reflection), or between multiple-back reflection echoes, on a standardized base line of the A-scan display. The base line of the A-scan display should be adjusted for the desired thickness increments. 7.1.2 Flaw detectors with numeric readout are a combination pulse ultrasound flaw detection instrument with an A-scan display and additional circuitry that provides digital thickness information. The material thickness can be electronically measured and presented on a digital readout. The A-scan display provides a check on the validity of the electronic measurement by revealing measurement variables, such as internal discontinuities, or echo-strength variations, which might result in inaccurate readings. 7.1.3 Thickness readout instruments are modified versions of the pulse-echo instrument. The elapsed time between the initial pulse and the first echo or between multiple echoes is converted into a meter or digital readout. The instruments are designed for measurement and direct numerical readout of specific ranges of thickness and materials. 8.2 Case II—Delay Line Single-Element Search Unit: 8.2.1 Conditions—When using this search unit, it is necessary that the equipment be capable of correcting for the time during which the sound passes through the delay line so that the end of the delay can be made to coincide with zero thickness. This requires a so-called “delay” control in the instrument or automatic electronic sensing of zero thickness. 8.2.2 In most instruments, if the material standardize circuit was previously adjusted for a given material velocity, the delay control should be adjusted until a correct thickness reading is obtained on the instrument. However, if the instrument must be completely standardized with the delay line search unit, the following technique is recommended: 8.2.2.1 Use at least two standardization blocks. One should have a thickness near the maximum of the range to be measured and the other block near the minimum thickness. For convenience, it is desirable that the thickness should be “round numbers” so that the difference between them also has a convenient “round number” value. 8.2.2.2 Place the search unit sequentially on one and then the other block, and obtain both readings. The difference between these two readings should be calculated. If the reading thickness difference is less than the actual thickness difference, place the search unit on the thicker specimen, and adjust the material standardize control to expand the thickness range. If the reading thickness difference is greater than the actual thickness difference, place the search unit on the thicker specimen, and adjust the material standardize control to decrease the thickness range. A certain amount of over correction 7.2 Search Units—Most pulse-echo type search units (straight-beam contact, delay line, and dual element) are applicable if flaw detector instruments are used. If a thickness readout instrument has the capability to read thin sections, a highly damped, high-frequency search unit is generally used. High-frequency (10 MHz or higher) delay line search units are generally required for thicknesses less than about 0.6 mm [0.025 in.]. Measurements of materials at high temperatures require search units specially designed for the application. When dual element search units are used, their inherent nonlinearity usually requires special corrections for thin sec- 656 ASME BPVC.V-2021 ARTICLE 23, SE-797/SE-797M (a) Proportional sound path increases with decrease in thickness. (b) Typical reading error values. FIG. 2 Dual Transducer Nonlinearity thickness end of the range. The variation is also shown schematically in Fig. 2(a). Typical error values are shown in Fig. 2(b). 8.3.2 If measurements are to be made over a very limited range near the thin end of the scale, it is possible to standardize the instrument with the technique in Case II using appropriate thin standardization blocks. This will produce a correction curve that is approximately correct over that limited range. Note that it will be substantially in error at thicker measurements. 8.3.3 If a wide range of thicknesses is to be measured, it may be more suitable to standardize as in Case II using standardization blocks at the high end of the range and perhaps halfway toward the low end. Following this, empirical corrections can be established for the very thin end of the range. 8.3.4 For a direct-reading panel-type meter display, it is convenient to build these corrections into the display as a nonlinear function. is usually recommended. Reposition the search unit sequentially on both blocks, and note the reading differences while making additional appropriate corrections. When the reading thickness differential equals the actual thickness differential, the material thickness range is correctly adjusted. A single adjustment of the delay control should then permit correct readings at both the high and low end of the thickness range. 8.2.3 An alternative technique for delay line search units is a variation of that described in 8.2.2. A series of sequential adjustments are made, using the “delay” control to provide correct readings on the thinner standardization block and the “range” control to correct the readings on the thicker block. Moderate over-correction is sometimes useful. When both readings are “correct” the instrument is adjusted properly. 8.3 Case III—Dual Search Units: 8.3.1 The method described in 8.2 (Case II) is also suitable for equipment using dual search units in the thicker ranges, above 3 mm [0.125 in.]. However, below those values there is an inherent error due to the Vee path that the sound beam travels. The transit time is no longer linearly proportional to thickness, and the condition deteriorates toward the low 8.4 Case IV—Thick Sections: 8.4.1 Conditions—For use when a high degree of accuracy is required for thick sections. 657 ARTICLE 23, SE-797/SE-797M ASME BPVC.V-2021 from steel walls having elevated temperatures is high (too thick) by a factor of about 1 % per 55°C [100°F]. Thus, if the instrument was standardized on a piece of similar material at 20°C [68°F], and if the reading was obtained with a surface temperature of 460°C [860°F], the apparent reading should be reduced by 8 %. This correction is an average one for many types of steel. Other corrections would have to be determined empirically for other materials. 8.4.2 Direct contact search unit and initial pulse synchronization are used. The display start is delayed as described in 8.4.4. All display elements should be linear. Incremental thickness is displayed on the A-scan display. 8.4.3 Basic standardization of the sweep will be made as described in Case I. The standardization block chosen for this standardization should have a thickness that will permit standardizing the full-sweep distance to adequate accuracy, that is, about 10 mm [0.4 in.] or 25 mm [1.0 in.] full scale. 8.4.4 After basic standardization, the sweep must be delayed. For instance, if the nominal part thickness is expected to be from 50 to 60 mm [2.0 to 2.4 in.], and the basic standardization block is 10 mm [0.4 in.], and the incremental thickness displayed will also be from 50 to 60 mm [2.0 to 2.4 in.], the following steps are required. Adjust the delay control so that the fifth back echo of the basic standardization block, equivalent to 50 mm [2.0 in.], is aligned with the 0 reference on the A-scan display. The sixth back echo should then occur at the right edge of the standardized sweep. 8.4.5 This standardization can be checked on a known block of the approximate total thickness. 8.4.6 The reading obtained on the unknown specimen must be added to the value delayed off screen. For example, if the reading is 4 mm [0.16 in.], the total thickness will be 54 mm [2.16 in.]. 9.6 Instrument—Time-base linearity is required so that a change in the thickness of material will produce a corresponding change of indicated thickness. If a CRT is used as a readout, its horizontal linearity can be checked by using Practice E317. 9.7 Back Reflection Wavetrain—Direct-thickness readout instruments read the thickness at the first half cycle of the wavetrain that exceeds a set amplitude and a fixed time. If the amplitude of the back reflection from the measured material is different from the amplitude of the back reflection from the standardization blocks, the thickness readout may read to a different half cycle in the wavetrain, thereby producing an error. This may be reduced by: 9.7.1 Using reference blocks having attenuation characteristics equal to those in the measured material or adjusting back reflection amplitude to be equal for both the standardizing blocks and measured material. 9.7.2 Using an instrument with automatic gain control to produce a constant amplitude back reflection. 9. Technical Hazards 9.1 Dual search units may also be used effectively with rough surface conditions. In this case, only the first returned echo, such as from the bottom of a pit, is used in the measurement. Generally, a localized scanning search is made to detect the minimum remaining wall. 9.8 Readouts—A-scan displays are recommended where reflecting surfaces are rough, pitted, or corroded. 9.8.1 Direct-thickness readout, without an A-scan display, presents hazards of misadjustment and misreading under certain test conditions, especially thin sections, rough corroded surfaces, and rapidly changing thickness ranges. 9.2 Material Properties—The instrument should be standardized on a material having the same acoustic velocity and attenuation as the material to be measured. Where possible, standardization should be confirmed by direct dimensional measurement of the material to be examined. 9.9 Reference Standards—Greater accuracy can be obtained when the equipment is standardized on areas of known thickness of the material to be measured. 9.3 Scanning—The maximum speed of scanning should be stated in the procedure. Material conditions, type of equipment, and operator capabilities may require slower scanning. 9.10 Variations in echo signal strength may produce an error equivalent to one or more half-cycles of the RF frequency, dependent on instrumentation characteristics. 9.4 Geometry: 9.4.1 Highest accuracy can be obtained from materials with parallel or concentric surfaces. In many cases, it is possible to obtain measurements from materials with nonparallel surfaces. However, the accuracy of the reading may be limited and the reading obtained is generally that of the thinnest portion of the section being interrogated by the sound beam at a given instant. 9.4.2 Relatively small-diameter curves often require special techniques and equipment. When small diameters are to be measured, special procedures including additional specimens may be required to ensure accuracy of setup and readout. 10. Procedure Requirements 10.1 In developing the detailed procedure, the following items should be considered: 10.1.1 Instrument manufacturer’s operating instructions 10.1.2 Scope of materials/objects to be measured 10.1.3 Applicability, accuracy requirements 10.1.4 Definitions 10.1.5 Requirements 10.1.5.1 Personnel 10.1.5.2 Equipment 10.1.5.3 Procedure qualification 10.1.5.4 Training or certification levels 10.1.6 Procedure 10.1.6.1 Measurement conditions 10.1.6.2 Surface preparation and couplant 10.1.6.3 Standardization and allowable tolerances 10.1.6.4 Scanning parameters 9.5 High-temperature materials, up to about 540°C [1000°F], can be measured with specially designed instruments with high-temperature compensation, search unit assemblies, and couplants. Normalization of apparent thickness readings for elevated temperatures is required. A rule of thumb often used is as follows: The apparent thickness reading obtained 658 ASME BPVC.V-2021 10.1.7 Report 10.1.7.1 Procedure used 10.1.7.2 Standardization record 10.1.7.3 Measurement record ARTICLE 23, SE-797/SE-797M 11.1.1.3 Size, frequency, and type of search unit. 11.1.1.4 Scanning method. 11.1.2 Results. 11.1.2.1 Maximum and minimum thickness measurements. 11.1.2.2 Location of measurements. 11.1.3 Personnel data, certification level. 11. Report 11.1 Record the following information at the time of the measurements and include it in the report: 11.1.1 Examination procedure. 11.1.1.1 Type of instrument. 11.1.1.2 Standardization blocks, size and material type. 12. Keywords 12.1 contact examination; nondestructive testing; pulseecho; thickness measurement; ultrasonics APPENDIX (Nonmandatory Information) X1. Typical Multi-Step Thickness Gage Reference Blocks TABLE OF DIMENSIONS Metric Block 4A, mm U.S. Customary Block, in. Metric Block 4B, mm Legend Dimension Tolerance Dimension Tolerance Dimension Tolerance T1 T2 T3 T4 L W 0.250 0.500 0.750 1.000 0.75 0.75 0.001 0.001 0.001 0.001 0.02 0.05 6.25 12.50 18.75 25.00 20.0 20.0 0.02 0.02 0.02 0.02 0.5 1.0 5.00 10.00 15.00 20.00 20.0 20.0 0.02 0.02 0.02 0.02 0.5 1.0 NOTE 1—Material to be as specified. NOTE 2—Surface finish: “T” faces Ra 0.8 µm [32 µin.] max. Other surfaces Ra 1.6 µm [63 µin.] max. NOTE 3—Location for optional 1.5 mm [1⁄16 in.] diameter through hole used for block support during plating; center 1.5 mm [1⁄16 in.] from block edges. NOTE 4—All “T” dimensions to be after any required plating or anodizing. NOTE 5—In order to prevent sharp edges, minimize plating buildup, or remove in-service nicks and burrs, block edges may be smoothed by beveling or rounding, provided that the corner treatment does not reduce the edge dimension by more than 0.5 mm [0.020 in.]. FIG. X1.1 Typical Four-Step Thickness Reference Blocks 659 ARTICLE 23, SE-797/SE-797M ASME BPVC.V-2021 TABLE OF DIMENSIONS Metric Block 5A, mm U.S. Customary Block, in. Metric Block 5B, mm Legend Dimension Tolerance Dimension Tolerance Dimension Tolerance T1 T2 T3 T4 T5 L W 0.100 0.200 0.300 0.400 0.500 0.75 0.75 0.001 0.001 0.001 0.001 0.001 0.02 0.05 2.50 5.00 7.50 10.00 12.50 20.0 20.0 0.02 0.02 0.02 0.02 0.02 0.5 1.0 2.00 4.00 6.00 8.00 10.00 20.00 20.00 0.02 0.02 0.02 0.02 0.02 0.5 1.0 NOTE 1—Material to be as specified. NOTE 2—Surface finish: “T” faces Ra 0.8 µm [32 µin.] max. Other surfaces Ra 1.6 µm [63 µin.] max. NOTE 3—Location for optional 1.5 mm [1⁄16 in.] diameter through hole used for block support during plating; center 1.5 mm [1⁄16 in.] from block edges. NOTE 4—All “T” dimensions to be after any required plating or anodizing. NOTE 5—In order to prevent sharp edges, minimize plating buildup, or remove in-service nicks and burrs, block edges may be smoothed by beveling or rounding, provided that the corner treatment does not reduce the edge dimension by more than 0.5 mm [0.020 in.]. FIG. X1.2 Typical Five-Step Thickness Reference Blocks 660