Uploaded by Tzy Woei Yap

MA4871 CA2 YAP TZY WOEI

advertisement
Chapter 1: Specimen and defects
The specimen used is the titanium grade 5 (Ti-5, also known as Ti-6Al-4V) block that’s cut into a custom
shape. The dimension and the defects are indicated as below on Figure 1 & 2.
Defect A
Defect B
60mm
30mm
60mm
100mm
120mm
Z
120mm
X
Y
Isometric view of specimen with defects
Figure 1: Dimensions of Titanium Grade 5 (Ti-5) specimen with Defects
12mm
Transverse corner
crack
10mm
12mm
8mm
Jagged Transverse
Subsurface Crack
Jagged Transverse
Surface Crack
12mm
14mm
8mm
Jagged Transverse
Subsurface Crack
Side view of specimen with dimensions
Jagged Longitudinal
Subsurface Crack
3mm Dia.
Spherical void
Side view of specimen with defects
Figure 2: Dimensions & location of defects
There are two types of defects in the specimen, namely Defect A and Defect B. Defect A is a corner
surface cracks and subsurface defects that appears on the XY and XZ plane, while Defect B is a surface
and subsurface defect on the XY and XZ plane with a 3mm diameter of spherical void located inside
the specimen. These defects consist of cracks with a width of 1mm. The suitability of different NonDestructive Testing (NDT) methods for detecting these defects will be discussed. The NDT methods
under consideration are:
1.
2.
3.
4.
5.
Ultrasonic Testing (UT) for sub surface defects
Radiographic Testing (RT) for sub surface defects
Liquid Penetrant Testing (LP) for surface defects
Magnetic Particle Testing (MPT) for surface defects
Eddy Current Testing (ECT) for surface defects
Chapter 2: Defect A
Defect A is a surface corner defects characterized by two jagged transverse cracks on the XY and XZ
plane respectively. Two jagged subsurface cracks penetrate into the specimen to a maximum depth of
10mm in the form of a transverse crack, as illustrated in Figure 2. The lengths of the surface transverse
cracks are measured at 12mm and 10mm, respectively. The first subsurface transverse crack extends
downwards to a depth of 2mm. As the crack progresses in the Z direction, the depth increases to 6mm,
and then reduces to 4mm further along. This crack with the depth of 10mm is now connected to the
second subsurface jagged transverse crack that grow in the Y direction.
The surface corner cracks in Defect A have a width of 1mm and a length of 10mm in Z-axis and 12mm
in Y-axis, which makes them not visible to the naked eye due to their small size. Furthermore, the
extent of penetration of these cracks cannot be observed visually. To detect Defect A, Liquid Penetrant
(LP) testing can be performed. The titanium grade 5 (Ti-5) which is grey in colour would benefit from
the use of a red penetrant to achieve good contrast. Prior to performing Fluorescent Penetrant (FP)
testing, surface preparation is necessary to clean the surface. LP can be first applied on the top planes
of the specimen. Once capillary action has occurred, a developer can be added. Excess developer can
then be wiped away to reveal the longitudinal cracks. Deeper cracks would exhibit a higher volume of
penetrant, allowing for comparison of crack depth. However, the exact depth of the crack may not be
determined using this method.
Another NDT to be considered is Magnetic Particle Testing (MPT). The defect surface should be
thoroughly cleaned to remove any dirt, grease, oil, paint, or other contaminants that could interfere
with the inspection process. The surface should be cleaned using suitable methods such as degreasing,
wiping, sandblasting, or chemical cleaning, depending on the specific requirements and condition of
the material. After cleaning, the surface should be prepared to ensure good contact between the
magnetic particles and the surface. This typically involves roughening the surface to create a suitable
profile for the magnetic particles to adhere to. Methods such as grinding, sanding, or shot blasting can
be used to roughen the surface, taking care not to damage the material. Magnetic particles, which are
usually a fine powder consisting of ferromagnetic materials, are applied to the prepared surface of
specimen. This can be done using a dry method where the particles are dusted or sprayed onto the
surface, or a wet method where the particles are suspended in a liquid carrier and applied to the
surface by spraying or immersion. After the magnetic particles are applied, a magnetic field is applied
to the component. This can be done using a permanent magnet, an electromagnet, or a magnetic yoke.
The magnetic field induces magnetic flux into the material, and any magnetic particles that are
attracted to areas of flux leakage caused by defects, such as cracks or discontinuities, will accumulate
and form a visible indication on the surface. Any indications of magnetic particles that have
accumulated at areas of flux leakage will be observed. The indications can be interpreted to identify
the presence, location, size, and shape of defects. However, MPI could not detect the specific depth of
the defects.
On the other hand, Eddy Current Testing (ECT) can be performed to identify the defects. To perform
ECT, it is preferable to have already located the defects, ideally using LP testing. EC testing should be
conducted after LP testing. To calibrate the EC machine, a reference specimen with similar defects
should be produced, allowing for calibration based on the skin depth formula to determine the
required frequency for defect detection. The frequency selected should match the largest depth of the
defects, which is 12mm, in order to detect all the cracks. ECT probes should be applied along the
located surface cracks to obtain a response. Since Defect A is primarily a surface defect, the phase
angle of the response is expected to be similar. However, the response curve of the ECT probe would
vary in size due to the differences in total depth and size of each flaw being measured. An issue with
using ECT for Defect A is that it originates from three planes and runs through an edge, which may
affect the accuracy of ECT measurements due to the presence of open spaces and multiple planes that
can impact the formation of eddy currents and, consequently, the results. Additional settings may be
needed for the ECT machine to account for measuring edges and differences in specimen height.
Another limitation of EC is that it can only accurately measure straight vertical cracks and may not
accurately measure the total length of a jagged crack, potentially only capturing the vertical length,
leading to inaccurate measurements.
Among the three NDT methods mentioned above, namely LP, MPT, and ECT, LP and ECT are more
suitable for detecting defect A. Both methods require preparation and time, as well as post-processes
after conducting the inspection. For instance, ECT necessitates demagnetization of the specimen, while
LP requires surface cleaning to remove the dyes. Furthermore, only ECT has the capability to measure
and obtain the depth and size of Defect A, whereas LP can only detect the surface portion of Defect A.
In fact, a combination of both LP and ECT would be favourable, with LP being used initially for crack
detection, which would expedite the process, followed by ECT for relative measurements of crack
depth and size.
Chapter 3: Defect B
Defect B has the surface and subsurface defect on the XY and XZ planes with a 3mm diameter of
spherical void located inside the specimen. It is located 60mm from the YZ plane in X direction and
60mm from XZ plane in Y direction. The cracks with 1mm width and propagate in the X, Y and Z
direction. To identify and quantify the defect, both Ultrasonic Testing (UT) and Radiographic Testing
(RT) were taken into consideration, while thermography was ruled out due to the material being a
titanium, which would result in rapid heat dissipation to the surroundings. There is no specific
constraint on the direction in which UT and radiography can be applied. In this section, we will explore
the application of both UT and radiography from the top XY plane and on the YZ plane from the size.
UT involves utilizing soundwaves generated by piezoelectric transducers to identify flaws. UT
equipment typically consists of a pulser, receiver, signal capture, and waveform display. It is essential
to verify if the UT equipment is capable of measuring or detecting the specific defect.
Firstly, the near field length is determined using the formula 𝑧𝑔= 𝐷2/4𝛌. The measured 𝑧𝑔 must be at
least 20mm, as that is the depth of the nearest flaw. Additionally, the beam divergence must be wide
enough to capture the entire crack and the back surface. UT can be applied from either the top or the
side plane, as it only requires access from one side. However, it is important that the attenuation
caused by the specimen is low enough to ensure accurate results. A couplant, which is a substance
that improves the transmission of ultrasound, should be applied to the surface where the transducer
is placed, as there is a significant impedance mismatch between air, specimen, and transducer. A
specific gate can be used to collect the crack displayed at the required surface. For instance, either a
singular transducer or an ultrasonic array can be applied for the measurement.
When applying UT from the top surface of the specimen, an angular beam inspection would be
preferable. Depending on the specific measurement requirements, we could use a single transducer
in either pulse mode or a pitch catch mode. Pulse mode would allow UT to detect cracks based on the
reflection of soundwaves from the defects which provide a result of variations in amplitude. However,
for this method to be effective, the cracks would need to be perpendicular to the UT wave produced.
In the case of Defect B, which has cracks in multiple orientations, the angles of these flaws would need
to be determined beforehand in order to detect them accurately. By adjusting the angle of Perspex or
the material to match the angle of the individual cracks, it is possible to generate a UT wave that is
perpendicular to the flaws. This can be achieved using Snell's law, 𝑠𝑖𝑛𝜃1/𝑐1= 𝑠𝑖𝑛𝜃2/𝑐2 which states that
the sine of the angle of incidence in one medium is equal to the sine of the angle of incidence in
another medium, multiplied by the respective velocities of the two media. By considering the
material's velocity (c) and the angle of the Perspex (𝜃1), it is possible to calculate 𝜃2, which would result
in a UT wave that is perpendicular to the angled cracks. To ensure detection of all the cracks,
soundwaves would need to be applied in multiple directions. The cracks could be visualized in different
settings, such as A-scan, B-scan, or C-scan, which may show the raw data or converted images. However,
these methods may not fully capture the defects as they are composed of multiple cracks with varying
orientations.
Another UT technique that can be used is Time-of-Flight Diffraction (TOFD), which operates in pitch
catch mode and is independent of flaw orientation. This method produces a D-scan, which is an image
captured from a top view, obtained from an A-scan that indicates the upper and lower tips of
reflections. However, UT may not be able to detect defects that are aligned with each other. For
instance, the longitudinal crack in Defect B would likely be detected as it is located below multiple
cracks coming from different directions. UT would only detect cracks that are located above the
longitudinal crack. Furthermore, jagged cracks may not be accurately represented in the UT images.
While soundwaves can still reflect from these jagged cracks, the reflected soundwave may not directly
reach the receiver, resulting in inaccurate readings. These voids can cause the directed soundwave to
reflect in multiple directions, which may prevent the soundwave from being reflected back to the
receiver and thus not displaying a reading.
We can apply UT from a different direction which is from the right side of the YZ plane, producing
waves in the X direction. As 2 of the defects do not propagate in X direction, those defects would be
oriented perpendicular to the waves propagating in the X direction. In order to ensure proper
transmission of ultrasound, a coupler would still need to be applied to the side plane. Alternatively, a
contact transducer that produces waves in the X direction could be used. To account for the differing
thickness of the specimen in X direction, it is preferable to set a gate width of 60mm. This gate would
encompass the location of Defect B, while also mitigating potential confusion caused by varying
specimen thickness. A C-scan which is displaying the length in both Y and Z directions would produce
the entire flaw of defect B.
Another approach Radiographic Testing (RT) can be applied to examine Defect B. Access is required
from both sides of the material for Radiography, as X-rays need to be applied from one side and the
film needs to be applied on the other side. If the X-rays are applied from the top plane, a film must be
put on the bottom plane. X-rays are directed towards the block along the line of the suspected defects.
The X-ray source is positioned at the appropriate distance and angle, and the exposure settings.
Defects would appear as dark images on the film, while the rest of the specimen would produce lighter
images. In the case of Defect B, it is necessary for it to have a contrast that is greater than 2% of the
total depth in the X and Z direction. This means that cracks or voids with a depth of 1.4mm or greater
in the X and Z direction would be detectable. A low kilovolt (kV) X-ray would be selected to obtain a
contrasting and accurate image of Defect B. However, the selected energy must be sufficient to
penetrate the full depth of the specimen, which is 60mm in Z direction and 120mm in X direction.
Using appropriate exposure factors, such as lower kilovolt (kV) and milliampere (mA) settings, can help
optimize image quality and sharpness and resulting in a contrasting image. Overexposure or
underexposure can affect image sharpness. The distance between the X-ray source and the film can
enhance the geometric sharpness. However, the SFD should still be within the recommended range to
ensure proper exposure and avoid image distortion. The X-ray beam should be directed at an angle
parallel to the cracks in defect B in order to obtain an accurate portrayal. This would necessitate
multiple X-ray shots to effectively capture the defects. Initially, a shot would be taken vertically from
the top plane to measure the depth of the defects, which would include spherical voids and potentially
some of the angled cracks. Additionally, six additional shots would be required at different angles to
capture all the remaining cracks. When observed from a top view, the results would reveal a single
long crack in the X and Y direction because the defects are situated in the same XY plane, and
radiography in this direction would not be able to distinguish between individual cracks. Additionally,
cracks located at different depths but on the same XY plane would also remain undetected. The X-ray
applied directly would only be able to detect the void with 3mm diameter and the longitudinal cracks
that is propagated in X direction. As a result, only the cracks on the topmost layer would be shown and
the radiography method would not yield precise measurements of the depth of each defect. Instead,
it would only provide an estimation based on the coloration displayed on the film, with deeper cracks
appearing darker. Radiography testing can be performed from the right side of the specimen which is
YZ plane and the film to be placed on the left side of the specimen to fully detect the Defect B. The
voids with diameter of 3mm and 2 of the transverse jagged cracks can be detected.
Comparing the UT and RT, UT would be preferred NDT because it has the capability to fully capture
Defect B when measured from the side view. Both methods necessitate training and in-depth
knowledge for proper operation. UT may be considered a more cost-effective technique compared to
radiography. Besides, UT would fully display Defect B, including voids when applied directly from the
side view. However, UT may require more time compared to radiography as UT requires additional
calibration and prerequisites, such as accurate orientation of flaws in Defect B, and higher energy levels
than radiography. On the other hand, radiography would provide a quicker indication of the presence
of defects, albeit with less accuracy. Despite the X-rays not being parallel to the defects, radiography
would still be able to detect the defects, albeit with diminished contrast. As a result, radiography would
be a more straightforward and expedient choice, although it may have limitations in terms of accuracy.
Chapter 4: Conclusion
The LP and MPI methods used together are the best NDTs to measure surface defects. UT is preferred,
but radiography is necessary to fully detect and measure Defect B for subsurface defects. In fact, these
methods rely on the availability of information about defect dimensions and locations, which may not
always be provided in real-world scenarios, NDT would be required to detect and characterize each
defect, which can be time-consuming and costly. Combining different NDT techniques may result in
more precise defect detection, but it could also increase costs, which may not always be preferred.
Additionally, real-world defects are often more complex than the defects discussed above, making
their detection even more challenging.
Download