The University of Texas Permian Basin College of Engineering Department of Mechanical Engineering Ehsanul Kabir, Ph.D MENG-4205 Thermo-Fluid Lab Impact Test November 30, 2020 Table of Contents Abstract ........................................................................................................................................................................ 2 Introduction .................................................................................................................................................................. 2 Methodology ................................................................................................................................................................. 3 Theory ....................................................................................................................................................................... 3 Apparatus/Equipment/Materials ................................................................................................................................ 5 Figure 1: The Charpy and Izod Impact Testing Unit (EEICI) [2] ........................................................................ 5 Figure 2: The Protractor Point [2]........................................................................................................................ 6 Figure 3: Charpy and Izod Hammers [2] ............................................................................................................. 6 Figure 4: Supporting clamps [2] .......................................................................................................................... 7 Figure 5: Placement of plastics on supporting clamps [2].................................................................................... 7 Procedure .................................................................................................................................................................. 8 Results ........................................................................................................................................................................... 9 Table 1: Data collected while performing the Izod Test ...................................................................................... 9 Table 2: Data collected while performing the Charpy Test .................................................................................. 9 Table 3: Results calculated by performing the Izod Test ..................................................................................... 9 Table 4: Results calculated by performing the Charpy Test ................................................................................. 9 Figure 6: White material after impact on the Izod Test...................................................................................... 10 Figure 7: Blue material after impact on the Izod Test ........................................................................................ 10 Figure 8: Clear material after impact on the Izod Test ....................................................................................... 10 Figure 9: White material after impact on the Charpy Test ................................................................................. 11 Figure 10: Blue material after impact on the Charpy Test ................................................................................. 11 Figure 11: Clear material after impact on the Charpy Test ................................................................................ 11 Discussion ................................................................................................................................................................... 15 Conclusion .................................................................................................................................................................. 15 References ................................................................................................................................................................... 17 Appendix..................................................................................................................................................................... 18 1 Abstract The purpose of this experiment is to observe materials toughness and indirectly its ductility. To do so, this experiment utilizes the Charpy and Impact testing unit (EEICI) from Edibon, that is equipped with a pendulum supported on bearings, that can be fixed with either a Charpy or Izod hammer, a protractor that indicates the angle and three plastics (white, blue, clear) . For both the Charpy and the Izod, begin the experiment by placing the pendulum on the releasing mechanism with a starting angle of 150°. Release once and collect the first angle from the protractor, then release again from 150° to collect the second angle. Repeat three times for both hammers. Once completed, run the experiment by placing a plastic onto the supporting clamps and release the pendulum from 150, then collect the angle given by the protractor. This is repeated for both hammers and each plastic. Analyzing the data shows that the clear plastic absorbed the most energy (5.49 Joules) in both methods of testing, while the blue material absorbed the least amount of energy (1.09 Joules) during the Charpy Test and absorbed the same amount of energy (1.90 Joules) as the white material during the Izod Test. Introduction The purpose of this experiment is to test the impact strength of each of the three materials being observed. There are two testing methods used, the Izod Impact Test and the Charpy Impact Test. The two methods vary only by the position and orientation in which the sample is secured prior to impact. The primary goal is to calculate the amount of potential energy when the pendulum is all the way up then calculate the amount of energy remaining after it has struck the sample. These two values may then be subtracted to find the amount of work or energy it takes to break each sample which corresponds to the impact strength of each material, respectively. 2 This method of impact testing is incredibly important to the world in which we live today. While these tests may not always replicate the exact situation that a material will experience once in use, they provide practical modes of failure, and values for the impact strength of a material which can then be used in numerous ways. The most important information gathered from these tests is how the material fails. This is critical to the use of certain materials in the design of structures. For example, a brittle material which fails almost instantly due to one crack propagating the entire sample upon impact, would never want to be used in a structural application. If a structure begins to fail it should do so gradually, like a ductile material so that everyone has the ability to get out and then proceed to make repairs rather than collapsing immediately like a brittle material which could injure or kill the occupants. Methodology Theory: The Izod and Charpy Tests are both forms of impact tests that are used to determine the strength of materials. They both work similarly, in that they both utilize a hammer attached to a pendulum to break materials. However, the difference lies in the way that the material specimens are held. During the Izod test the specimen is placed vertically and clamped down on the bottom end and is placed so that the face with the notch faces the direction that the hammer is coming from. During the Charpy Test the specimen is placed horizontally so the notched side is facing away from the direction that the hammer comes from. In both impact tests the initial angle every time is α = 150°. Using the initial angle (α), the initial energy (𝐸!) can be found as: 𝐸! = 𝑚 × 𝑔 × 𝑙 × '1 − 𝑐𝑜𝑠(𝛼)0 3 (1) Where, 𝑚 is the mass of the hammer, 𝑔 is gravity (𝑔 = 9.81 " ), #! and 𝑙 is the length of the pendulum. Both devices have their own constant, initial energy. Now, the final energy (𝐸$) can be found by using a similar equation to equation 1: 𝐸$ = 𝑚 × 𝑔 × 𝑙 × '1 − 𝑐𝑜𝑠(𝛽)0 (2) Where, 𝛽 is the angle given by the indicator, 𝑚 is the mass of the hammer, 𝑔 is gravity (𝑔 = 9.81 "!), and 𝑙 is the length of the pendulum. Next, the absorbed energy (𝐸%) can be found # by using the initial energy (𝐸!) and final energy (𝐸$) as follows: 𝐸% = 𝐸! − 𝐸$ (3) Now, the indicator loss (𝑊&') can be found by: 𝑊&' = 𝑊( − 𝑊) (4) The indicator loss is found by running the impact tests without specimens three different times on each device. 𝑊) is found by using the angle shown by the indicator as it is pushed from 0° to where the pendulum stops. To find 𝑊( the indicator is left where the pendulum stopped to find 𝑊). The angle that is displayed after the pendulum has pushed the indicator further is used to find 𝑊(. Both 𝑊( and 𝑊) utilize equation 2 to be calculated. The total indicator loss is calculated three times for each test and then averaged together to find one total indicator loss (𝑊&' +,+) for each test: 𝑊&' +,+ = 4 -"#$ .-"#! .-"#% / (5) Where, 𝑊&'( is first calculated indicator loss, 𝑊&') is the second calculated indicator loss, and 𝑊&'/ is the third calculated indicator loss. The energy due to friction losses (𝐸01) on both devices can be found utilizing the total indicator loss (𝑊&' +,+) and the initial energy (𝐸!), found on their respective devices: 𝐸01 = 𝐸! − 𝑊&' +,+ (6) Apparatus/Equipment/Materials: Figure 1: The Charpy and Izod Impact Testing Unit (EEICI) [2] Figure 1 displays the Charpy and Izod Impact Testing Unit (EEICI). The EEICI is composed of a pendulum supported on bearings with an arm of 330mm long and a releasing mechanism that’s set at the initial position of the test (150°). The EEICI has a protective clear 5 cover that shields the user when conducting the experiment. The EEICI also has a protractor pointer that measures the angle after impact from the pendulum. Figure 2: The Protractor Point [2] Figure 2 displays the Protractor Point that measures the angle after impact from the pendulum. Figure 3: Charpy and Izod Hammers [2] Figure 3 displays the Charpy and Izod hammers utilized in this experiment. The left image is the Charpy and the right image is the Izod. 6 Figure 4: Supporting clamps [2] Figure 4 displays the supporting clamps used for the plastics uses. The image on the left is used for the Charpy experiment and the image on the right is used for the Izod experiment. Figure 5 below will display how the plastics are placed onto the supporting clamps. Figure 5: Placement of plastics on supporting clamps [2] Figure 5 displays how the plastics will be placed on the supporting clamps. It is important to note that this is how the experiment will be conducted. 7 Procedure: It is important to note that the experiment for the Charpy and Izod hammers will be conducted the same with the exception of the placement of the plastics on the supporting clamps. To begin the experiment, place the pendulum on the releasing mechanism with a starting angle of 150° and set the protracting pointer at an angle of 0°. Without placing anything on the supporting clamps, lower the protective clear shield and release the pendulum. Once the pendulum is done moving, obtain the angle indicated by the protracting pointer. Without moving the protractor pointer back to zero, place the pendulum back on the releasing mechanism and lower the protecting shield then release the pendulum. Once the pendulum comes to a stop, obtain the angle indicated by the protracting pointer. Repeat this three times and for both the Charpy and Izod hammers. After completing the previous steps for each hammer, refer to Figure 5 to continue the experiment for each of the plastics provided. Place one of the plastics onto the supporting clamps, refer to figure 5 for proper placement, and set the pendulum at the starting position of 150° on the releasing mechanism. Lower the protective shield and release the pendulum. Once the pendulum comes to a stop, observe the fracture of the plastic, and obtain the angle indicated by the protracting pointer. Repeat this step for each of the plastics for the Charpy and Izod hammers. 8 Results Table 1: Data collected while performing the Izod Test Table 2: Data collected while performing the Charpy Test Table 3: Results calculated by performing the Izod Test Table 4: Results calculated by performing the Charpy Test 9 Figure 6: White material after impact on the Izod Test Figure 7: Blue material after impact on the Izod Test Figure 8: Clear material after impact on the Izod Test 10 Figure 9: White material after impact on the Charpy Test Figure 10: Blue material after impact on the Charpy Test Figure 11: Clear material after impact on the Charpy Test 11 Tables 1 and 2 organize the data that was collected while performing the Izod and Charpy Tests, respectfully, on the EEICI unit. Both methods of testing were initially performed without a specimen present 3 times, to determine how friction from the indicator effects the energy of the devices. The first angle listed in the sections with no material listed represents the angle that the device achieved with friction and the second value represents the angle achieved without friction. Once, the friction losses observations were complete, 3 different specimens were inserted separately into both devices to observe the energy absorbed by each specimen during impact. Tables 3 and 4 display the values calculated from performing Izod and Charpy Tests. Both tables show that the clear material absorbed the most energy, or is the most ductile, in both methods of testing. While the blue material absorbed the least amount of energy during the Charpy Test and absorbed the same amount of energy as the white material during the Izod Test. Figures 6 through 8 show the way each material fractured while performing the Izod Test. Figure 6 shows the white material, Figure 7 shows the blue material, and Figure 8 shows the clear material. The types of fractures formed during the Izod Test show that each material is ductile. When the clear material was tested, the specimen did not fracture until the pendulum swung back down. This shows that the clear material is the most ductile out of the three specimens. Figures 9 through 11 show the way each material fractured while performing the Charpy Test. Figure 9 shows the white material, Figure 10 shows the blue material, and Figure 11 shows the clear material. These figures confirm what was observed during the Izod Test; all the materials are ductile while the clear material has the highest ductility. The clear material having the highest ductility is proven by the fact that it did not fully fracture. This means that it absorbed the highest amount of energy, which makes it the most ductile. 12 Sample Calculations: First, looking at Table 3. Equation 1 can be used to find the initial energy of the pendulum: 𝐸! = 𝑚 × 𝑔 × 𝑙 × '1 − 𝑐𝑜𝑠(𝛼)0 𝐸! = 1.425 × 9.81 × 0.33 × '1 − 𝑐𝑜𝑠(150)0 𝐸! = 8.608259757 𝐽 Next, the final energy of the white material in Table 3 can be found using equation 2: 𝐸$ = 𝑚 × 𝑔 × 𝑙 × '1 − 𝑐𝑜𝑠(𝛽)0 𝐸$ = 1.425 × 9.81 × 0.33 × '1 − 𝑐𝑜𝑠(117)0 𝐸$ = 6.707479909 𝐽 Now, the energy absorbed by the white material in Table 3 can be found using equation 3: 𝐸% = 𝐸! − 𝐸$ 𝐸% = 8.608259757 − 6.707479909 𝐸% = 1.900779848 𝐽 13 Now, looking at Table 1 the indicator loss can be found three different times using equation 4: 𝑊&' = 𝑊( − 𝑊) 𝑊&'( = '1.425 ∗ 9.81 ∗ 0.33 ∗ (1 − cos(128))0 − '1.425 ∗ 9.81 ∗ 0.33 ∗ (1 − cos(123))0 𝑊&'( = 0.327637 𝐽 𝑊&') = '1.425 ∗ 9.81 ∗ 0.33 ∗ (1 − cos(130))0 − '1.425 ∗ 9.81 ∗ 0.33 ∗ (1 − cos(122))0 𝑊&') = 0.520679 𝐽 𝑊&'/ = '1.425 ∗ 9.81 ∗ 0.33 ∗ (1 − cos(130))0 − '1.425 ∗ 9.81 ∗ 0.33 ∗ (1 − cos(122))0 𝑊&'/ = 0.520679 𝐽 Next, the average of the indicator losses must be found using equation 5: 𝑊&' +,+ = 𝑊&' +,+ = 𝑊&'( + 𝑊&') + 𝑊&'/ 3 0.327637 + 0.520679 + 0.520679 3 𝑊&' +,+ = 0.456332 𝐽 14 Referring to Table 3 the energy due to friction losses can be found using equation 6: 𝐸01 = 𝐸! − 𝑊&' +,+ 𝐸01 = 8.608259757 − 0.456332 𝐸01 = 8.151928047 𝐽 Discussion No discussion questions were provided for the experiment. Some sources of possible error in this experiment could have included mechanical and human errors. Some mechanical errors that could have happened during this experiment include the EEICI machine not working the way it was intended to. The EEICI machine could have not been calibrated right, with something malfunctioning in a way that it could have not let it get calibrated. This could have also been a result of human errors as calibration is conducted and monitored by humans. Some other human errors could include the arm of the EEICI machine not being put on the correct angle that it needed to fully break some of the more ductile samples. A way this experiment could be improved is to have a more automated EEICI machine where human interaction is not needed as much. The hinged cover door of the machine could be automated to where less humans need to be involved as to improve results and reduce errors. Conclusion In conclusion, the results for the Izod test on Table 1 show that the final angle of the white and blue test samples ended with the same angle, at 117 degrees. The last sample, the clear one, ended with the lower final angle at 71 degrees. 15 Table 2 shows the results of the Charpy test on the test samples. The blue test sample had the largest final angle at 118 degrees, while the white stopped at 109 degrees and the clear one had the smallest final angle at 55 degrees. Tables 3 and 4 show the results of the calculations of the initial energy and the energy due to losses of the Izod and Charpy tests respectively. The Izod test had a larger initial energy than the Charpy test with 8.608259757 J. The difference of the initial energies is 3 J. The energy due to friction losses was also larger for the Izod test at 8.17928047 J with the difference between the tests being approximately 4 J. The test samples with two biggest differences between them were the white and blue test samples for the Izod test. 16 References [1] EDIBON. “EEICI.” Charpy and Izod Impact Testing Unit, 2016, www.edibon.com/EQUIPOS/EEICI/EEICI_EN.pdf. [2] EDIBON EEICI unit of Charpy and Izod Impact Testing Practice Manual 17 Appendix 18