Proper Microscopic Observation of Animal Cells and Plant Cells under Light Microscope Sharlimar Faith Calimutan Science, Technology, Engineering, and Mathematics Strand – A Janiuay National Comprehensive High School Abstract Observing cells involves refined practice and equipment such as microscopes, and thus, can be a difficult task to do. Hence, this report aimed to learn and apply the proper methods for operating the optical microscope to observe and compare animal cells and cheek cells. To acquire this, the student has done laboratory work utilizing a 10x magnification eyepiece, iodine solution for staining and a light microscope focusing on a low power objective (LPO) lens and a highpower objective (HPO) lens to examine cheek cells and onion cells as representations of animal cells and plant cells, respectively. Results have found the distinguishing features between cheek cells and onion cells. Under LPO and HPO, onion cells have prominent cell walls, cytosol, and nucleus, while cheek cells have an irregular shape, cytoplasm, and a slightly noticeable cell membrane. However, the image of cheek cells did not emphasize the nucleus of the cells under LPO and HPO. This report suggests that there is a further need to improve the calibration of the microscope to attain a better view of cheek cells and other cells as well as consider and improve the factors involving the matter. Keywords: Animal Cells, High Power Objective Lens, Low Power Objective Lens, Magnification, Plant Cells Introduction Microscopy involves the technical practice of viewing samples and objects using microscopes that are either too small for the unaided eye to perceive or are too small to be viewed clearly by the human eye (What Is Microscopy? | The University of Edinburgh, 2018). There are many types of microscopes but for this activity, light microscope was used. It is the type of microscopy in which light is transmitted from a source that is on the sample's other side from the objective lens. To focus the light on the sample and attain optimum brightness, it is typically passed through a condenser. Following the sample, the light passes via the objective lens, which enlarges the image of the sample, and then to the oculars, where the larger picture is viewed. There are two types of light microscopes based on the nature of the light source: the simple microscope and compound microscope. Since it is light microscopes are cheaper than electron microscopes, it was used in simple scientific laboratories to improve the scientific knowledge of the students. Scientific laboratory is essential to the holistic development of a learner as it involves multiple tasks to understand complex concepts. Therefore, schools and governments are encouraged to build scientific laboratories in their countries to be utilized by students and improve their scientific innovation. In the case of the Philippines, there is a further need to enhance the scientific innovation and skills of the students given several studies and reports discovering the lack of scientific knowledge of Filipino students. One such study is the 2019 edition of the Trends in International Mathematics and Science Study (TIMSS), which involves Grade 4 Filipino students (Kelly, Centurino, Martin, & Mullis, 2020). The study gave the Philippines scores of 297 and 249 in mathematics and science, respectively, which is the lowest among the 58 countries involved in the study. Looking at the situation, Filipino students need to observe, experiment, and explore concepts like scientists, and to do that, a scientific laboratory is vital. In addition, research by American Chemical Society (ACS) has stated that students who engage in well-designed laboratory experiences develop problem-solving and criticalthinking skills, as well as gain exposure to reactions, materials, and equipment in a lab setting. And given this activity of observing cells, participants will be able to improve their scientific proficiency and literacy both individually and interactively. Thus, this report aimed not only to accurately calibrate the microscope and apply approaches to observing cells but also to assess the scientific capability of the learner in laboratory work. Materials and Methods The main purpose of this report is to determine how to use and adjust the microscope from low power objective to high power objective during the observation of animal cells and plant cells. Thus, this activity evolved from enhancing the skills in proper calibration or focusing of the cells. Before starting the experiment, preparation of the specimens and other materials such as laboratory tools, including the microscope, toothpicks, iodine solution, glass sides, and droppers, as well as the cleanliness of the environment has been done. First, the students proceeded to start with onion cells by peeling a thin skin of an onion epidermis and putting it on the center of the glass slide. Next, a small amount of iodine stain was put on the top of the specimen using the dropper and waited for about 10 seconds so the stain could penetrate the cells before a clear glass cover was put on top and repaired the air bubbles. Then the specimen was put on the center stage of the microscope and held tightly using the stage clip. Afterward, the experimenter adjusted the microscope, starting with LPO, further calibrating the course adjustment knob and fine adjustment knob and improving the light to be reflected to enhance the resolution of the specimen when viewed under the microscope. Calibration continued for minutes until the experimenter found a detailed and clear view of the specimen. After observing under LPO, the student switched the objective to HPO to magnify the specimen again. Lastly, the observer documented the specimen for this report. Before observing cheek cells, the observer cleaned the glass slide and glass cover again. Then the student obtained the cheek cells by gently scraping the inside of the mouth with the use of a toothpick. And just like the onion cells, the cheek cells were put on top and in the center of the glass slide, then stained with iodine solution before being put underneath the coverslip. With the same procedure, calibration of the microscope was done until better images of the cheek cells under LPO and HPO were achieved. Figure 1. (a) Onion cells under LPO; (b) Onion cells under HPO; The arrow designated the parts of the cell. In the second experiment, as seen in Figure 2, most of the cheek cells under LPO and HPO do not have a noticeable nucleus even if the specimen was stained with iodine solution, but have cell membrane, cytosol, and shape irregularities. The expected result was to have a small dark-regioned nucleus and clarity of the image. Results and Discussion In the first experiment, onion cells were observed to have an evident nucleus, cell wall, cell membrane, and cytoplasm. In Figure 1., the prominent parts of the onion cells under LPO and HPO have been shown. The rectangular boundary depicts the cell wall and inside the cell is the cytoplasm, while the slightly circular opaque matter on each cell is the nucleus. It can be learned that iodine solution enhances the resolution of the cells as well as the nucleus is not clear under LPO but became visible under HPO. Figure 2. (a) Cheek cells under LPO; (b) Cheek cells under HPO This may have involved interferences such as improper calibration, contrasting, and illumination errors. There is too much dust and debris on the surface, as well as the cells are in low concentration. It can also be observed that there is resolution, contrasting and focusing errors that have occurred during the experimentation. In lieu of the experiment, the errors in the images were assessed. First, poor contrast can be a usual indication that the substage condenser aperture diaphragm is opened too wide, causing flare and considerably reducing image contrast. This problem plagues all types of photomicrography, including color, black & white, and digital. To improve the contrast of the specimen in a light microscope, adjust both the field diaphragm and the aperture diaphragm. (Abramowitz & Davidson). Next, poor resolution of the specimen can be the result of improper adjustment of the substage condenser aperture diaphragm. When a microscope is properly configured for Köhler illumination (the method of providing the optimum specimen illumination), the aperture diaphragm should have an iris diameter opening that lies between 65 and 80 percent of the objective aperture. Hence, the condenser aperture diaphragm should be set to a position that will provide a compromise mixture of direct and deviated light that depends, to a large degree, on the absorption, diffraction, and refraction characteristics of the specimen. (Abramowitz & Davidson). Meanwhile, contamination of dust can appear as colored artifacts when using color film, and as black or gray spots on black and white film just like on the specimen. If the debris appears in sharp focus, then it is usually occurring from contamination of a lens surface residing in a plane conjugate to the focused specimen. The field lens and any glass (often color balancing or correcting filters) adjacent to the microscope field diaphragm are often prone to collecting dirt, and these components lie near one of the principal image conjugate planes. Contamination near the field diaphragm will appear in sharp focus on film, and scratches or dirt on filters placed near the field diaphragm are often a source of haze and debris artifacts. Glass surfaces that have been scratched will appear as unsharp areas on photomicrographs, as will bubbles embedded in the lamp condenser glass or in heat-absorbing filters. (Abramowitz & Davidson). Another common source of dirt and debris is the specimen itself, which also resides in a conjugate plane. Dust adhering to the surface of the coverslip will appear as unfocused dark spots on the film, while contamination underneath the coverslip near or mixed with the specimen will usually appear as debris that is in focus with the specimen. (Abramowitz & Davidson) Lastly, the lack of proper focus and/or blurry images represent one of the most common errors in photomicrography. The source of these errors is usually the result of vibration in the microscope stand or improper adjustment of the focal distance between the optics and the film plane. Often, the image will appear in sharp focus through the eyepieces, but resulting photomicrographs are blurred or unsharp. This is an indication that the film plane and viewing optics may not be parfocal. (Abramowitz & Davidson) Conclusion The present report discusses the importance of laboratory work in describing animal cells and plant cells. The results indicated the parts of the cell in each specimen under LPO and HPO. However, the parts of the animal cells were not clearly evident in the experiment. Hence, several errors in the problem were evaluated and explained. Factors affecting the image quality of the cells, such as poor contrast, blurry images, dust and debris, and poor resolution as a result of improper calibration and preparation, were described. Therefore, there is a further need to consider and limit those factors to avoid the problem when doing experimental work. Acknowledgment This work was accomplished with the support of the General Biology I teacher of Science, Technology, Engineering, and Mathematics Strand-A (STEM-A) named Jay Cee Jondonero, the supervisor of the said strand, who is named Digna Esteva Jolito, as well as the parents giving the researcher the financial support to buy the needed materials. Other students have also helped the researcher by interacting and comparing their own findings to achieve better results for the experiment. References Abramowitz, M., & Davidson, M. W. (n.d.). Troubleshooting microscope configuration and other common errors. Photomicrography Microscope Configuration and Other Common Errors | Olympus LS. Retrieved September 11, 2022, from https://www.olympuslifescience.com/en/microscoperesource/primer/photomicrography/errors/ Importance of Hands-on Laboratory Science - American Chemical Society. (n.d.). American Chemical Society. Retrieved September 11, 2022, from https://www.acs.org/content/acs/en/policy/publicp olicies/education/computersimulations.html Kelly, D.L., Centurino, V.A.S., Martin, M.O., & Mullis, I.V.S. (Eds.) (2020). TIMSS 2019 Encyclopedia: Education Policy and Curriculum in Mathematics and Science. Retrieved from Boston College, TIMSS & PIRLS International Study Center website: https://timssandpirls.bc.edu/timss2019/encycloped ia/ What is Microscopy? | The University of Edinburgh. (2018, September 27). The University of Edinburgh. https://www.ed.ac.uk/clinicalsciences/edinburgh-imaging/for-patients-studyparticipants/tell-me-more-about-my-scan/whatis-microscopy