TSFE-X01010 SHIMADZU MEDICAL SYSTEMS NISHINIHON CORPORATION TECHNICAL SUPPORT & SERVICES DEPT. TSFE-X01010 - ii - TSFE-X01010 CONTENTS 1. Introduction ......................................................................................................1 1.1. 1.2. 2. 3. Outline .....................................................................................................................................1 Reference ...............................................................................................................................2 Cautions ............................................................................................................3 Features of MUX-10 .........................................................................................5 3.1. 3.2. 3.3. 3.4. 3.5. 3.6. 4. Super Maintenance Function .................................................................................................5 FVR Adjustment......................................................................................................................5 Starter ......................................................................................................................................5 Measured Value Display Function..........................................................................................5 TMA Waveform .......................................................................................................................5 Exposure Counter/Error Log Display Function ......................................................................5 Periodical Inspection ........................................................................................6 4.1. 4.2. 4.3. 4.4. 5. 6. 7. 8. Mechanical Inspection Items ..................................................................................................6 Electrical Inspection Items ......................................................................................................7 Inspection Items Related to X-ray Tube.................................................................................7 Cleaning ..................................................................................................................................7 Periodically Replaced Parts..............................................................................8 Recommended Spare Parts ..............................................................................9 PARTS MAP ................................................................................................... 11 REPLACING THE PARTS AND ADJUSTING............................................14 8.1. REPLACING THE BRAKE SHOE.......................................................................................14 8.2. Adjusting the brake ...............................................................................................................15 8.3. Replacing the caster..............................................................................................................16 8.4. Replacing the X-ray tube. .....................................................................................................17 8.4.1. Replacing the collimator. ............................................................................................17 8.4.2. Removing the X-ray tube............................................................................................18 8.4.3. Reassemble X-ray tube unit.......................................................................................20 8.4.4. Procedure after exchange x-ray tube.........................................................................23 8.4.5. Adjusting after replacing the X-ray tube. .................................................................24 8.5. Adjusting of tube rotation. .....................................................................................................24 8.6. Adjusting of arm catcher .......................................................................................................25 8.7. Arm bearing adjusting. ..........................................................................................................25 8.8. Replacing the wire.................................................................................................................26 8.9. Replacement of Collimator Lamp.........................................................................................28 8.10. Checking IGBT..................................................................................................................29 8.11. Waveform Confirmation and Adjustment.........................................................................30 9. Troubleshooting ..............................................................................................31 9.1. 9.2. 9.3. When Power Does Not Turn On or Off ................................................................................31 When X-ray Tube Rotation Sound Is Abnormally Large .....................................................31 When Collimator Lamp Does Not Turn On or Off................................................................32 - iii - TSFE-X01010 9.4. When Lock Cannot Be Released or Set..............................................................................33 9.5. UD125P-CH Self-diagnosis Function...................................................................................34 9.5.1. Alarm message list......................................................................................................34 9.5.2. Error message list......................................................................................................34 9.6. Countermeasures against Error Messages .....................................................................35 9.6.1. H1 and H2 ...................................................................................................................35 9.6.2. H3..................................................................................................................................35 9.6.3. F01 and F02 ................................................................................................................35 9.6.4. F03 ................................................................................................................................36 9.6.5. F11 (KV OVER)..........................................................................................................36 9.6.6. F12 (KV UNDER)......................................................................................................37 9.6.7. F13 (NO KV) ...............................................................................................................38 9.6.8. F14 (KV BRAKE).......................................................................................................39 9.6.9. F21 (MA OVER) and F22 (MA UNDER) ..............................................................40 9.6.10. F23 (NO MA) ..........................................................................................................41 9.6.11. F31 (IF OVER) and F32 (IF UNDER)...............................................................42 9.6.12. F33 (NO IF).............................................................................................................43 9.6.13. F41 ............................................................................................................................44 9.6.14. F43 ............................................................................................................................45 9.6.15. F44 and F45 ............................................................................................................46 9.6.16. F46 ............................................................................................................................47 9.6.17. F61 ............................................................................................................................48 9.6.18. F62 ............................................................................................................................49 9.6.19. F6E ...........................................................................................................................49 9.6.20. F6F ...........................................................................................................................49 10. 11. MUX-10 Block Diagram.................................................................................50 Explanation on Each PCB..............................................................................52 11.1. 11.2. 11.3. 11.4. 11.5. 11.6. 11.7. NEXSC-MUX10................................................................................................................52 X CONT-2000 PCB.........................................................................................................54 INVERTER UNIT MUX...................................................................................................57 INPUT PCH-99 ................................................................................................................58 SHEET PANEL ................................................................................................................59 MUX AEC I/F ...................................................................................................................60 RC RECEIVER ................................................................................................................61 - iv - TSFE-X01010 1. Introduction 1.1. Outline This service manual describes the manual type Mobile X-ray Unit MUX-10. Use this service manual as the reference data of service works for inspection, repair, etc. -1- TSFE-X01010 1.2. Reference The following reference are related to this service manual. All of these reference are required in service works. Make sure to use them together with this service manual. The information described in the following reference is not described in this service manual in principle. ① MUX-10 Installation Manual Document No.: M503-E308 This manual describes the information required in installation and maintenance as well as detailed adjustment methods. When replacing PCBs, make sure to perform adjustment using the described adjustment methods. ② MUX-10 Connection Diagram Book The document No. is not given. This manual shows circuit diagrams of the MUX-10. ③ MUX-10 Parts List Document No.: M503-4003 This list shows the information required in ordering parts. ④ MUX-10 Operation Manual Document No.: M503-E006 This manual describes the specifications, the operating procedures and the information on maintenance and inspection, guarantee, disposal, etc. -2- TSFE-X01010 2. Cautions Make sure to read thoroughly the cautions below before starting a work during installation, inspection or maintenance. Fixing the Counterweight Make sure to fix the counterweight inside the column to the column with bolt before removing the X-ray tube and the collimator. If the counterweight is not fixed, the MUX-10 may become unbalanced and extremely dangerous. Caution on Waveform Adjustment Make sure to take the following action to prevent charging before observing the IGBT gate waveform. If the IGBT gate waveform is observed in two positions at the same time in the charged status, an accident may be caused. Remove the fuse F4 from the INPUT PCH-99 PCB. Make short-circuit the CP3 (CHECK1) and the CP4 (CHECK2) on the X CONT PCB. Caution on Power ON Do not turn on the power while the extension NEXSC PCB is provided. If such a PCB is provided, the MUX-10 may malfunction. Caution on High Voltage Generation Do not generate high voltage while the high voltage cable and the X-ray tube are not connected. -3- TSFE-X01010 Prevention of Electrical Shock (1) Turn OFF the power before starting maintenance of the inside of the cabinet. Even if the power is turned OFF, electric charge may remain for a while. Make sure that the LD1 on the INPUT PCH-99 PCB is lit off before starting maintenance. Prevention of Electrical Shock (2) After disconnecting the high voltage cable, make sure to let the high voltage plug pin be in contact with the grounded portion so that the plug pin is discharged. Prevention of Charge When touching a PCB, make sure to wear the wrist band for prevention of static electricity so that charge can be prevented. Protection against Radiation When checking the filament of the X-ray tube or visually checking discharge inside the X-ray tube, make sure to take proper protective action such as use of lead glass. -4- TSFE-X01010 3. Features of MUX-10 The features of the MUX-10 regarded as important to service are described below. 3.1. Super Maintenance Function The initial setting mode and the adjustment mode are available without setting the DIP switches on the NEXSC PCB. Refer to Appendix D in the Installation Manual. 3.2. FVR Adjustment The MUX-10 offers two adjustment modes, the “two-point adjustment mode” in which the CPU calculates other data after the operator has adjusted only two points in the FVR table and the conventional “all-point adjustment mode” in which the operator adjusts all data in the FVR table. Perform adjustment in the two-point adjustment mode in installation and adjustment, then confirm the FVR in the all-point adjustment mode. 3.3. Starter The starter circuit is assembled in the INVERTER UNIT PCB. The starter starts in 1.6 sec at 125V, and is driven intermittently at every 10 sec in the steady state. 3.4. Measured Value Display Function Exposure for 32 msec or more is required to use the measured value display function. Have this fact in mind in adjusting the FVR. 3.5. TMA Waveform The TMA waveform is of 1 V/50 mA. In conventional units, the preset TMA value was displayed for 10 msec after exposure was started. In the MUX-10, the preset value is not displayed. 3.6. Exposure Counter/Error Log Display Function The MUX-10 has the function to count the number of times of exposure and to restore the error log. Refer to Appendix B in the Installation Manual. -5- TSFE-X01010 4. Periodical Inspection Perform the periodical inspection in accordance with the tables below. 4.1. Mechanical Inspection Items Inspection item 1 2 3 4 5 6 7 Are bolts tightened securely in the tube holder? Are flat countersunk head screws in the collimator mounting area and flush head bolts in the X-ray tube and the mounting ring tightened Tighten the bolts. Apply LOCKTITE. securely? (Remove the collimator for check.) Is the column fixed securely? Are the fixing bolts tightened securely? Are front wheel casters fixed securely? Does the X-ray tube rotate smoothly without abnormal sound? Is the tension of two wire ropes equivalent? Are the double-nuts tightened securely? Are the wire ropes free from rust? Are the wire ropes free from flaws? Are element wires free from cuts and fine splits? Does the X-ray tube move up and down smoothly 9 without abnormal sound? Does the column rotate smoothly without 10 abnormal sound? 8 11 Countermeasures Is the X-ray tube expanded smoothly without abnormal sound? 12 Does the lock release switch operate normally? Apply the lubricant. Adjust the wire nuts. Apply the lubricant. Replace the wire ropes. Clean the rails. Apply the lubricant. Or release the lock, and check for defects. Remove the cover from the arm end face, and check the movement visually. Investigate the cause in accordance with connection diagrams. 13 Is the grip fixed securely? Tighten the bolts. 14 Is the travel brake released and locked normally? Adjust the gap. Clean the wheels. 15 Does the collimator lamp light normally? Replace the lamp. 16 Are cables not caught? Treat the cables again. 17 Does the arm catch unit operate normally? Check the movement visually. -6- Inspection result TSFE-X01010 4.2. Electrical Inspection Items Inspection item 1 2 3 4 5 6 7 8 4.3. Countermeasures Measure the supply voltage at each check pin. Observe actually the TKV and TMA waveforms to check whether the tube voltage, the tube current, the exposure time and the tube current time product are output in accordance with the settings. Confirm the photograph density if the phototimer is provided. Adjust the DC power output. (Adjustment is available only for 24 VDC.) Adjust again the FVR and IGBT gate waveforms. Adjust the phototimer again. Is the high voltage cable plug tightened securely? Tighten it securely. Are terminals and connectors fixed or contacted securely? Do the panel display and the sheet keys function normally? Are cables free from breakage? Check for breakage in the cable reel plug especially. Fix the terminals and the connectors. Check contact of the connectors. Or replace the panel. Is the ground belt free from damages? Replace the ground belt. Replace the cables. Inspection Items Related to X-ray Tube Inspection item Countermeasures 1 Is the X-ray tube rotation sound normal? Replace the X-ray tube. 2 Is the X-ray tube free from oil leak? Replace the X-ray tube. 4.4. Inspection result Inspection result Cleaning Make sure to clean the MUX-10 in the periodical inspection. Cleaning position 1 Appearance of MUX-10 2 Inside of MUX-10 3 Wheels, brake face and electromagnetic face of lock and rail. Caution Remove dirt with cloth soaked in neutral detergent .And use carefully so that the MUX-10 is not damaged. Remove dusts by using a vacuum cleaner. Fire may be caused by dusts and foreign substances. Remove dirt. -7- Result TSFE-X01010 5. Periodically Replaced Parts Replace the following parts in the periodical inspection. Part Name Part No. Used Area Q’ty Halogen lamp 062-65003-02 Collimator 1 LITHIUM BATTERY ASSY 501-78646 NEXSC 1 FUSE, SC-10 072-01649-01 FUSE, 326010 072-01665-34 FUSE, 313002 072-01664-26 SILICONE RUBBER JACKET SILICONE WASHER 511-15061-13 X-RAY TUBE 2 SILICONE PACKING 532-23029 HIGH VOLTAGE TRANSFORMER 2 1 INPUT PCH-99 2 2 If the MUX-10 has been used for 10 years or more, replace the following parts. Part Name Part No. Product type Q’ty 503-46520-01 MUX-10 2 503-46520-04 MUX-10J 2 Wire rope -8- TSFE-X01010 6. Recommended Spare Parts The following parts are recommended as spare ones (including periodically replaced parts). RANK A QTY P/N DESCRIPTION 503-49045-01 062-65003-02 064-54801-98 064-54803-96 501-78646 072-06033-08 072-01664-20 072-02004-16 072-01649-01 072-01665-34 072-01664-26 074-80678-16 074-80678-86 502-21470 502-21572 502-21859 502-20775 060-39940-01 511-15061-13 CODE REEL ASSY LAMP HALOGEN #55939 SW PB A3AT-90A1-00L FLANGE A3A-222 LITHIUM BATTERY ASSY FUSE , 600FH-100 FUSE , 313001 FUSE , 218001 FUSE , SC-10 FUSE , 326010 FUSE , 313002 POWER SUPPLY , LDC60F2 POWER SUPPLY , LDA-50F-24-Y INVERTER UNIT MUX ASSY X CONT-2000 ASSY NEXSC-MUX10 INPUT PCH-99 ASSY IGBT , 2MBI300N-060 MAINTENANCE KIT MINI MUX-10 1 1 3 3 1 1 1 2 1 2 2 1 1 1 1 1 1 2 1 MUX-10J 1 1 3 3 1 1 1 2 1 2 2 1 1 1 1 1 1 2 1 RANK B P/N DESCRIPTION 503-34889-02 510-20036-05 503-51143 503-50495 503-51287 503-50093 511-64073 EARTH BELT PLUNGER PBF-12-KSN BRAKE SHOE MICRO SWITCH ASSY B MICRO SWITCH ASSY C RUBER CUSHION SHEET PANEL MUX-A QTY MUX-10 1 2 2 1 1 1 1 -9- MUX-10J 1 2 2 1 1 1 1 TSFE-X01010 RANK C QTY P/N DESCRIPTION 503-46520-01 503-46520-04 511-18026-01 511-37050 510-37081-04 502-20700-01 502-21479-01 065-90756-01 065-68009-48 065-40261-01 WIRE ROPE WIRE ROPE CABLE COVER WHEEL , 380 CASTER , 420SA H.T. TRANS D125PH-C1 H.T. CABLE MUX CIRCUIT PROTECTOR , CP22HS/15DB1 RELAY , G3NA-420B-DC5V CONTACTOR , SH5AC100V1A5B - 10 - MUX-10 2 1 2 2 1 1 1 1 1 MUX-10J 2 1 2 2 1 1 1 1 1 TSFE-X01010 7. PARTS MAP MAIN BODY RESISTANCE R1A,R1B,R2A,R2B SSR NEXSC DCPOWER SUPPLY PS1 DC POWER SUPPLY PS2 X CONT 2000 ARM SW SW4 INPUT PCH99 INVERTER UNIT T1TRNS TERMINAL CORDREEL - 11 - TSFE-X01010 CAPACITOR RELAY MG-D RESISTANCE RDA,RDB,RDC - 12 - TSFE-X01010 ELECTRIC LOCK TRANS T1 FOR REMOVING THE ELECTRIC LOCK EARTH BELT ELECTRIC LOCK - 13 - TSFE-X01010 8. REPLACING THE PARTS AND ADJUSTING 8.1. REPLACING THE BRAKE SHOE SPRING PLATE BRAKE COVER BRAKE BAR 1. Remove the brake cover. Remove the cassette box. 2. Remove the spring plate. Retighten the both bolts equally. 3. Remove the spring and the bolts. 4. Pull out the brake bar. 5. Remove the brake shoe assy from the brake bar. BRAKE SHOE SPRIG NUT BRAKE SHOE CASE SPRING PLATE - 14 - BRAKE BAR TSFE-X01010 8.2. Adjusting the brake Adjust the brake in case that brake release is not easy. Retighten the nut and rotate the eccentric shaft for changing the gap between the wheel and the brake shoe. After adjusting , tighten the nut securely. NUT ECCENTRIC SHAFT Apply the greese to below. grease - 15 - TSFE-X01010 8.3. Replacing the caster. The special tool is need to replace the caster. Spanner :503-37353 At first retighten the screw. Fix the shaft by special spanner and retighten the nut. remove the nut fix this shaft by spanner screw - 16 - TSFE-X01010 8.4. Replacing the X-ray tube. It is very dangerous to remove the X-ray tube and the collimator because the balance can not be keep. Fix the weight in the column by the bolt. 8.4.1. Replacing the collimator. Remove the collimator from the X-ray tube. Remove screws these 4 Remove the socket and high voltage cable from X-ray tube and remove the low voltage cable at the junction connector. Remove these screws - 17 - TSFE-X01010 8.4.2. Removing the X-ray tube. Remove the cable fixing at the arm and the column. Remove the U-nut at the end of the arm. remove the cover,U-nut,dics spring,detent washer,thrust washer Remove the X-ray tube assy. thrust washer detent washer U-nut disc spring - 18 - TSFE-X01010 X-ray tube assy U-nut,disc spring,detent washer,thrust washer - 19 - TSFE-X01010 8.4.3. Reassemble X-ray tube unit Remove the cover from the tube rotation unit attaching the tube to the arm, remove socket head cap screws (in 4 positions on each of the left and right sides) inside, then remove and move down the tube from the holding frame. CAUTION The weight of the tube ASSY is approximately 13 kg. Pay rigid attention in moving it down from the arm. Remove these socket head cap screws (in 4 positions). Remove these socket head cap screws (in 4 positions). Rotation unit on the right side Rotation unit on the left side Remove the left and right tube rotation units from the tube. As to the right side, loosen the U nut (new type lock nut not requiring a inner clip washer) in the center, then remove the rotation stopper, the spring washer, etc. After that, remove socket head cap screws from the shaft area, then remove the shaft area. As to the left side, remove the cover (countersunk head screws in 2 positions), then remove socket head cap screws from the inside. CAUTION Grease will be applied on sliding units at the end. Every time removing each part, clean the sliding units. Moreover, please be sure to decompose at the time of decomposition work, performing marking so that an attachment position and a direction may be known from back. It mistakes, and when assembling, it may not operate normally. Left rotation unit Remove these socket head cap screws (in 4 positions). Parts of left rotation unit - 20 - TSFE-X01010 Right rotation unit Before removing this rotation stopper, mark the attachment position. Remove this U nut. It is double-faced. Remove these socket head cap screws (in 4 positions). Parts of right rotation unit - 21 - TSFE-X01010 Remove the collimator rotation unit, and attach it to a new tube. Please be careful in the direction of stopper at this time. Remove these hexagon socket countersunk head screws (in 4 positions). Collimator rotation unit Attach the left and right rotation units to the new tube in the order of removal. At this time, clean the sliding units, and apply grease on the following portions before attachment. CAUTION Make sure to use grease not containing abrasive. Recommended grease: Albania grease 2 (Showa Shell) Apply grease on this portion. Apply grease on this portion. Apply grease on the entire this face. Apply grease on the entire this face. - 22 - Apply grease on this portion. TSFE-X01010 8.4.4. Procedure after exchange x-ray tube. At the end, attach the low voltage cable, the collimator and the high voltage plug, and treat the cables. At this time, attach the high voltage plug using the following procedure. ⅰ Make sure that the inside of the socket in the X-ray tube unit is not dirty. ⅱ Make sure that the plug and the ring nut are not dirty. (For the name of each part, refer to Fig. 6-31.) ⅲ Let the ring flange go through the ring nut. ⅳ Apply silicon oil on the silicon rubber jacket and the silicon washer. ⅴ Fit the silicon washer into the tip of the plug. ⅵ Put the plug into the socket. When the plug tip reaches the bottom of the socket, turn the plug to find the position in which the plug is fitted into the key way, then push the plug. ⅶ Tighten the ring nut with hand. ⅷ Tighten the rotation stopper screw so that the ring nut does not rotate. Ring nut Silicon washer Ring flange Silicon rubber jacket Not apply toa pin head. - 23 - High voltage cable Apply oil on this silicon rubber. Silicon oil TSFE-X01010 8.4.5. Adjusting after replacing the X-ray tube. Perform the aging. Perform FVR adjusting. 8.5. Adjusting of tube rotation. Forth for rotate the tube can be adjusted by U-nut. In case that rotation of X-ray tube is not smooth, please apply the greese refer to installation manual appendix F. - 24 - TSFE-X01010 8.6. Adjusting of arm catcher In arm catcher, there are two swiches. SW4 ARM SW SW4 should be ON when arm is hold. ARM switch should be ON when the lever is pulled. 8.7. Arm bearing adjusting. Adjust the eccentric bearing to decrease the gap of arm sliding. eccentric bearing - 25 - TSFE-X01010 8.8. Replacing the wire For replacing the wire, it need to fell down the column. Please perform the procedure referring to below. And take care to safty. Fix the weight in the column by the bolt. Remove the X-ray tube referring to “replacing the X-ray tube”. Remove the bolts at foot side of the column. fix the weight by the bolt Retighten the screw and remove the washer and pin. Pull out the shaft. Do not use the metal hammer, please use rubber hammer to protect the shaft from changing the shape. If shape of the end of the shaft is changed, it is impossible to assemble again. bearing with pully washer pin shaft pully collar - 26 - screw TSFE-X01010 Remove the bolt fixing the weight. Pull out the weight from the column. weight After remove the subweight and bearing mount, retighten and remove the metal fitting of wire. Remove the wire from the carrige. double nuts metal fitting In case of reassembling, do not forget to fix the weight by the bolt. - 27 - TSFE-X01010 8.9. Replacement of Collimator Lamp Replace the collimator lamp using the following procedure. CAUTION :Never touch directly the collimator lamp. If dirt is adhered on the lamp, the illuminance may be deteriorated. NOTE : After the power of the system has been OFF for a while or after a while after the key switch was set to OFF (that is, after the collimator lamp is cooled down), replace the collimator lamp. 1. 2. 3. 4. Remove the lamp outer cover of the collimator. Remove the lamp cover. Replace the lamp (P/N 062-65003-02 halogen lamp). Attach the covers again. Lamp outer cover (Remove screws in 3 positions.) Lamp cover (Remove screws in 2 positions.) - 28 - Replace the collimator lamp. (Never touch it directly.) TSFE-X01010 8.10. Checking IGBT In case that wave form of IGBT gate signal is abnormal, it means defective of IGBT. If you want to check IGBT only, refer to below data. This IGBT unit has two IGBT in one unit. Please refer to circuit diagram below. Reference value ( measured at prototype MUX-10) C2E1 + Digital meter − C2E1 + − Digital meter C1 terminal C1 − + Measured value 0.8MΩ infinity E2 − + 抵抗値 infinity 0.8MΩ E2 C2E1 IGBT circuit diagram - 29 - TSFE-X01010 8.11. Waveform Confirmation and Adjustment Refer to Appendix C in the MUX-10 Installation Manual. Make sure to take proper action such as short-circuiting in accordance with the instructions described there. - 30 - TSFE-X01010 9. Troubleshooting Read this chapter as reference of troubleshooting. 9.1. When Power Does Not Turn On or Off The outline of the power line is shown below. Check the line from the check pins on the PCB from power source and find the cause of the trouble. The main switch may be defective, the fuse may be blown out, or the DC power supply may be defective. 9.2. When X-ray Tube Rotation Sound Is Abnormally Large Prepare the noise meter, and measure the rotation sound. When the measured value is 65 dB or more at the distance of 1 m, the X-ray tube is abnormal. → Contact the Quality Assurance Department in Medical Systems Div. of Shimadzu Corporation. The rotation sound may be temporarily large in the early period (within 3 months after delivery), and may return to the normal sound after a while (due to sticking of the solid lubricant of the bearing). The skid time can be used to judge the rotation status of the X-ray tube. If the period of time from rotation at low speed to stop by inertia is 1 min or more, the X-ray tube is normal. - 31 - TSFE-X01010 9.3. When Collimator Lamp Does Not Turn On or Off The outline of the wiring related to the lamp is shown below. Check the wiring from the lamp terminals inside the collimator to power source, and find the cause of the trouble. The switch may be defective, or the fuse may be blown out, or the PCB may be defective,or the lump may be defective. T1-MUX-10 COLLIMATOR AC13V CN1 CN5 J135 JMU JT1IN F5 LAMP K6 INPUT PCH-99 SWITCH JINP JINP X CONT-2000 - 32 - NEXSC CIRCUIT PROTECTOR TSFE-X01010 9.4. When Lock Cannot Be Released or Set The outline of the wiring related to the locks is shown below. Check the wiring from the electromagnetic locks to DC power supply, and find the cause of the trouble. A switch may be defective, an electromagnetic lock may be defective, or the PCB may be defective. SL1 SL2 SL3 SL4 ARM SWITCH ARM SWITCH CN2 CN3 CN17 CN18 CN1 CN3 JMU DC24V K5A INPUT PCH-99 - 33 - TSFE-X01010 9.5. UD125P-CH Self-diagnosis Function 9.5.1. Alarm message list Display Description With the set exposure condition, the heat accumulated on the anode of the X-ray tube is over tolerance. Reduce the exposure condition, or stop exposure until warning code disappears. The heat accumulated on the anode of the X-ray tube has reached its tolerance limit. Stop H2 exposures until this warning code disappears. The temperature of X-ray tube assembly is too high. Stop exposures until this warning H4 code disappears. The battery for CPU memory backup needs replacing now. Contact your nearest E authorized Shimadzu service center. The life of battery for CPU memory backup has expired. Contact your nearest authorized EEE Shimadzu service center. NOTE) The CPU memory battery (lithium battery) backs up the SRAM on the NEXSC PCB. When the life of this lithium battery is expired, the contents displayed just before the power was turned off last time, the error log and the time setting are erased. Accordingly, the MUX-10 can be used though there is inconvenience. H1 9.5.2. Error message list For the action to be taken, refer to “7.16 counter measure against error message”. Display Description F01,F02 The supply voltage is out of its normal range. F03 The power supply circuit in the system is at fault. F11〜F14 The X-ray tube voltage control is at fault. F21〜F23 The X-ray tube current control is at fault. F31〜F33 The filament heating control is at foult. F41〜F46 The charge control is at fault. F61 The starter is at fault. F62 There is abnormality in the Bucky unit operation. F6E、F6F The irradiation control or photo timer - 34 - TSFE-X01010 9.6. Countermeasures against Error Messages Error messages displayed on the panel are described in detail, and the countermeasures are described below. 9.6.1. H1 and H2 The CPU calculates the accumulated heat capacity based on previous exposure, then issues these alarm messages. If an either of these messages is issued even while there is not so many exposure, check the voltage of lithium battery and initialize the CPU. 9.6.2. H3 When the thermal switch of the X-ray tube is activated, this message is issued. If this message is issued even while the X-ray tube is not hot at all, check the thermal switch signal. 9.6.3. F01 and F02 (detected by software) F01: The supply voltage is higher by approximately 20% or more than the normal value . F02: The supply voltage is lower by approximately 30% or more than the normal value . Check CP23 (LV) on X CON T -2000. W ithin range from 2.4 to 1.4V ? YES X CO NT or NEX SC is abnormal. → Replace them in turn. NO Fuse F5 on IN PU T PCH normal? NO Replace F5. YES Voltage between pins 3 and 5 of connector J135 from transformer T1-M UX10 13 V AC? YES NO T 1-M UX10 is defective. Or check power line to T 1M UX 10. - 35 - IN PU T PCH is abnormal. → Replace it. TSFE-X01010 9.6.4. F03 (detected by hardware) The -15V power supply for the control circuit is abnormal. Check -15V (CP26) on X CONT. Normal X CONT or NEXSC is abnormal. → Replace them in turn. Abnormal Check output of power supply PS2. Abnormal PS2 is abnormal. → Replace it. Normal -15V may be shortcircuited with GND on X CONT or other PCB. 9.6.5. F11 (KV OVER) (detected by software, at 20ms after exposure and each 100ms ) The measured tube voltage is higher by approximately 30 kV or more than the preset value. Measure TKV(CP9) and RKV(CP12) on X CONT. RKV abnormal? Or both TKV and RKV normal? YES Only TkV is abnormal. ① X CONT, ② INVERTER UNIT ③ High voltage transformer may be defective. - 36 - X CONT or NEXSC is abnormal. → Replace them in turn. TSFE-X01010 9.6.6. F12 (KV UNDER) The measured tube voltage is lower by 50% or more than the preset value (detected by software, at 20ms after exposure and each 100ms ) NOTE) It is most possible that glow discharge has occurred in the X-ray tube. When checking the IGBT gate waveform, take proper action such as short-circuiting to prevent generation of high voltage in accordance with Appendix C in the Installation Manual. Check TKV(CP9) and RKV(CP12) on X CONT. YES RKV abnormal? Or both normal? X CONT or NEXSC is abnormal. → Replace them in turn. NO Only either KVFB+ or KVFBoutput? YES One side of high voltage transformer is defective. →Replace high voltage transformer. NO FVR is too large. Adjust it. If it does not become normal, proceed to next step. Perform adjustment related to tube voltage shown in Appendix C in Installation Manual. If it does not become normal, proceed to next step. Abnormal Waveform at F(CP29), A/D(CP33) or B/C(CP34) normal? X CONT is abnormal. → Replace it. Normal IGBT gate waveform on INVERTER UNIT normal? Abnormal Normal IGBT single element check Normal Abnormal ① X ray tube ② High voltage transformer ③ High voltage cable may be defective. Replace IGBT. - 37 - INVERTER UNIT is abnormal. → Replace it. TSFE-X01010 9.6.7. F13 (NO KV) The measured tube voltage is approximately 10 kV or less (detected by software, at 20ms after exposure and each 100ms ) When checking the IGBT gate waveform, take proper action such as short-circuiting to prevent generation of high voltage in accordance with Appendix C in the Installation Manual. Fuse on INVERTER UNIT blown out? Blown out Normal Damaged IGBT damaged? Replace INVERTER UNIT. Or replace only IGBT and fuse. Check TKV(CP9) and RKV(CP12) on X CONT. Normal Replace fuse. RKV abnormal? Or both normal? YES X CONT or NEXSC is abnormal. → Replace them in turn. Only TKV is abnormal. Perform adjustment related to tube voltage shown in Appendix C in Installation Manual. Waveform Abnormal of F( CP29),A/D(CP33) or B/C(CP34) normal? X CONT is abnormal. → Replace it. Normal IGBT gate waveform on INVERTER UNIT normal? Abnormal Normal Normal IGBT single element check Abnormal High voltage transformer may be defective. Replace IGBT. - 38 - INVERTER UNIT is abnormal. → Replace it. TSFE-X01010 9.6.8. F14 (KV BRAKE) Abnormal measured tube voltage or X-ray tube discharge is detected (by the hardware). This error message actually postulates discharge in the high voltage system (especially the X-ray tube). This error may be eliminated by aging. Perform aging first. Perform aging. Same error occurs again. High voltage cable may be poorly contacted. Refer to "Appendix G High Voltage Cable Connection Procedure" in Installation M anual, and connect it again. Same error occurs again. ① X-ray tube, ② High voltage cable, ③ High voltage transformer, ④ X CONT or NEXSC may be defective. Judgement in the last item in the flow chart above is difficult. Abnormality in a PCB means that a false error message is issued, but the possibility is regarded as low. If an either of the following phenomena is detected, it can be clearly determined that discharge has occurred in the high voltage system. 1. Discharge sound “pop” is issued during exposure. 2. The error message is not issued at low kV, but is issued while the kV is rising. 3. The glass is clearly clouded (due to evaporation of the target material) when seen from the X-ray tube irradiation port. 4. The target is clearly rough when seen from the X-ray tube irradiation port. 5. Spark discharge like thick thunderbolt can be seen at the time of exposure from the X-ray tube irradiation port. (Make sure to take protective measures against radiation such as use of lead glass.) An aluminum filter is provided in the X-ray tube irradiation port. When checking the inside of the glass tube, remove the lead cone fast, then remove the aluminum filter also. - 39 - TSFE-X01010 9.6.9. F21 (MA OVER) and F22 (MA UNDER) (detected by software, at 20ms after exposure and each 100ms ) F21: The measured tube current is higher by approximately 50 mA or more than the preset value F22: The measured tube current is approximately 50% or less of the preset value . When these messages are displayed, check the heating system also in accordance with the flow chart for F31 and F32. Check RMA(CP13) and TMA(CP11) on X CONT. YES RMA abnormal? Or both normal? X CONT or NEXSC is abnormal. → Replace them in turn. Only TMA is abnormal. Adjust FVR again. Same error occurs again. X-ray tube may be defective. - 40 - TSFE-X01010 9.6.10. F23 (NO MA) The measured tube current is approximately 10 mA or less (detected by software, at 20ms after exposure and each 100ms ) Short-circuit bar (mA+ and mA-) attached to terminal block X1 on X CONT? NO Attach short-circuit bar. YES Check RMA(CP13) and TMA(CP11) on X CONT. RMA abnormal? Or both normal? YES X CONT or NEXSC PCB is abnormal. → Replace them in turn. Only TMA is abnormal. High voltage cable conductive? NO Replace high voltage cable. YES Breakage or poor contact in cable JFB between high voltage transformer and X CONT? YES Replace cable. NO ① High voltage transformer may be defective or ② X-ray tube filament may be broken. X-ray tube filament check method Remove the collimator. Look into the X-ray tube from the X-ray tube irradiation port, and check whether the filament is lit in the ready status. Make sure to use lead glass, etc. for protection against radiation. - 41 - TSFE-X01010 9.6.11. F31 (IF OVER) and F32 (IF UNDER) (detected by the software at each 800ms when AR! Income, and at each 400ms when ready up) F31: The measured filament current is higher by 50% or more than the preset value . F32: The measured filament current is lower by 50% or more than the preset value . Check FVR(CP15) and IFV(CP7) on X CONT. Both normal? Or only FVR abnormal? YES X CONT or NEXSC is abnormal. → Replace it. Only IFV is abnormal. NO C0 and CL waveforms normal? INVERTER UNIT or X CONT is abnormal. → Replace it. YES ① X CONT, ② INVERTER UNIT or ③ High voltage (filament) transformer may be defective. In case of F31, ② and ③ rarely occur. In the last item, ① and ② means defect in the detection system. Actually, the possibility of ③ is high. - 42 - TSFE-X01010 9.6.12. F33 (NO IF) The measured filament current is approximately 1 A or less . (detected by the software at each 800ms when AR! Income, and at each 400ms when ready up) YES Fuse F7 on INPUT PCH blown out? Replace F7. NO NO C0 and CL waveforms normal? Heating FET may be damaged. Refer to Appendix C in Installation Manual, and check CL and C0 waveforms. INVERTER UNIT or X CONT is abnormal. → Replace it. Waveform is abnormal. YES Replace INVERTER UNIT. Normal Check filament current of IFV(CP7) on X CONT. X CONT or NEXSC is abnormal. → Replace it. Abnormal High voltage cable may be poorly contacted. Refer to "Appendix G High Voltage Cable Connection Procedure" in Installation Manual, and connect it again. Exchange + and - sides of high voltage cable. Same error occurs again. Check X-ray tube filament visually. Filament is broken. Replace X-ray tube. Normal High voltage (filament) transformer may be defective. - 43 - TSFE-X01010 9.6.13. F41 The measured capacitor voltage is approximately 360V or more. The charging capacitor is charged beyond its withstand voltage. NO Transformer tap changed correctly for supply voltage? Change tap in accordance with Installation M anual. YES Perform "Appendix C Charging Voltage Adjustment" in Installation M anual. NO Adjustment performed correctly? INPUT PCH-99 is abnormal. → Replace it. Error message is displayed even after adjustment. NO SSR normal? Replace SSR. YES ① INPUT PCH-99, ② X CONT or ③ NEXSC may be defective. → Replace them in turn. - 44 - TSFE-X01010 9.6.14. F43 The measured capacitor voltage is approximately 20V or less at 3 sec after start of charging. The power may not be supplied. Transformer tap changed correctly for supply voltage? NO Change tap in accordance with Installation M anual. YES Fuse on INPUT PCH blown out? YES Replace fuse. NO Perform "Appendix C Charging Voltage Adjustment" in Installation M anual. NO Adjustment performed correctly? INPUT PCH-99 is abnormal. → Replace it. Error message is displayed even after adjustment. NO SSR normal? Replace SSR. ① INPUT PCH-99, ② X CON T or ③ NEXSC may be defective. → Replace them in turn. - 45 - TSFE-X01010 9.6.15. F44 and F45 F44: The measured capacitor voltage does not reach approximately 240 V (at which the charging control resistor is reset). F45: The measured capacitor voltage does not reach 310 V (which is the preset value). Transformer tap changed correctly for supply voltage? NO Change tap in accordance with Installation M anual. YES Perform "Appendix C Charging Voltage Adjustment" in Installation M anual. NO Adjustment performed correctly? INPUT PCH-99 is abnormal. → Replace it. Error message is displayed even after adjustment. ① INPUT PCH-99, ② X CONT or ③ NEXSC may be defective. → Replace them in turn. - 46 - TSFE-X01010 9.6.16. F46 The capacitor voltage rises while charging is stopped (detected by the software). The SSR may be short-circuited. NO SSR is abnormal. → Replace it. SSR normal? YES X CON T or NEXSC is abnormal. → Replace it. - 47 - TSFE-X01010 9.6.17. F61 Starter error (The starter does not issue the current detection confirmation signal.) The current does not flow in the stator coil, and the rotor does not rotate. NO SW1A set to ON on INVERTER UNIT? Set SW1A ON. YES F6 blown out on INPUT PCH? Voltage between CP7 and CP13,CP11 and CP13 should be 1V or more YES Replace F6. NO Check coil current of YC(CP7) and ZC(CP11) on INVERTER UNIT. Normal Replace X CONT or NEXSC. Abnormal X-ray tube coil resistance X-Y: 13±0.7W X-Z: 46±2.3 W X-ray tube connector pin layout Pin No. 1 = X Pin No. 2 = Y Pin No. 3 = Z Measure X-ray tube coil resistance. Abnormal Replace X-ray tube. - 48 - Normal Low voltage cable is poorly contacted. → Replace it. or INVERTER UNIT is abnormal. → Replace it. TSFE-X01010 9.6.18. F62 In 1 sec after HX1 is input, XOK is not sent back. This message is not displayed usually. Confirm the setting of the DIP switch. 9.6.19. F6E Exposure is not finished even after the backup time or beyond the preset mAs. The X CONT-2000 PCB (X-ray cutoff circuit) may be abnormal. 9.6.20. F6F The exposure control circuit is abnormal. In 250 msec after the EXR is output, the KVT (kV rising signal) is not sent back. This abnormality occurs when the shutdown signal IPOF is input at the same time with the EXR due to input of direct radiation during phototimer radiography. In other words, this abnormality cannot occur while the phototimer is not provided. If this message is displayed while the phototimer is not provided, the initial setting may be wrong or the X CONT-2000 PCB may be abnormal. - 49 - TSFE-X01010 10. MUX-10 Block Diagram INPUT PCH-99 CIRCUIT PROTECTOR LF POWER SOURCE JINP X CONT-2000 JINP NEXSC JM30 CHON MGRSON LV VP1 PRINTER J4 STON MGR-S CN1-1 VP1 JT1IN LV J135 TRANS T1-MUX10 CN1-2 JPS 13V JM2 J21 15V、5V JINV filament heating circuit VC to starter VC0 STOK TH1 TH2 V+ control resistance SL2 interface to sheet panel J9 AD,BC IGBT HAND SW AR1,HX1 RC RECEIVER X-RAY,READY,LAMPU T1,T2 TKV TMA STON JSSR starter DOOR CL REMOTE CN7 JREM charge control JFB SSRON CN8 JSP JM3 CHON JR12 SL3 SHEET PANEL MUX-A JSP V- MGRSON CN3 IGBT DRIVER CL,C0 CN2 SL1 AR1,HX1 FET DRIVE Q14,Q23 RSM1,RSM2 charege detection RELAY CIRCUIT TKV,TMA,IF JINV JSP 125V JMU KVBR5 TKV OVER -15VOK JM2 JSP STOK LV INVERTER UNIT-MUX 135V 280V 125V 100V CHON J22 MG-D J280 CN1 DC power supply PS1 CN2-2 THERMAL(TH1) J24 24V RECTIFIER 13V 135V ARM SW DC power supply PS2 CN2-1 CN3 CN15 ARM SW JAEC JSOUT CN7 KVFB+ KVFBN1 N2 X,Y,Z,TH1,TH2 SL4 COLLIMATOR CN15 X-ray tube H.V cable 13K,0K LAMPSW 0V - 50 - H.V trans DOOR1 DOOR2 REMOTE CONTROL CN4 SW4 ARM POSITION MUX AEC I/F PHM PH-REF EXR FSD2 IPOF PADDLE SIGNAL PADDLE TSFE-X01010 Description of major signals shown in the block diagram in the previous page Signal Name Description AD/BC IGBT driving signal AR1 Ready signal CHON Charging start signal CL,C0 Filament heating signal DOOR CL Arm accommodation confirmation signal EXR Exposure signal FSD2 Number of lighting fields selection signal HX1 Exposure IPOF Phototimer exposure shutdown signal KVBR5 Discharge detection signal KVFB+ + side of measured tube voltage signal KVFB- side of measured tube voltage signal LAMPU Collimator lighting signal from remote control unit LV Transformer T1 primary voltage confirmation signal MGRSON Charging suppression resistor reset signal N1,N2 Measured tube current signal PADDLE Voltage signal from phototimer detector SIGNAL PHM Phototimer selection signal PH-REF Phototimer reference signal Q14,Q23 Filament heating FET driving signal READY Ready signal from remote control unit RSM1,RSM2 Filament current signal STOK Starter confirmation signal STON Starter start signal TH1,TH2 X-ray tube thermal signal TKV Measured tube voltage signal TMA Measured tube current signal T1,T2 IGBT output signal VP1 Charging voltage detection voltage signal X-RAY Exposure switch signal from remote control unit -15VOK -15 VDC confirmation signal - 51 - TSFE-X01010 11. Explanation on Each PCB 11.1. NEXSC-MUX10 Function This is the CPU PCB which controls the MUX-10 using the software. Different from the NEXSC PCB used in other generators and fluoroscopy units, this NEXSC PCB is not equipped with an optical cable connector. (The NEXSC PCB equipped with an optical cable connector can be used in the MUX-10.) The following data is stored in this PCB. It is recommended to back up the following data in preparation for PCB replacement. Initial setting data FVR value Phototimer tube voltage correction value APR Number of times of exposure During PCB replacement When replacing the PCB, set the DIP switches first, then initialize the PCB. After that, perform the initial setting, adjust the FVR, adjust the phototimer, and register the APR. Lithium battery replacement When the life of the lithium battery is expired, the information (contents displayed just before the power was turned off last time, the error log and the time setting) stored in the SRAM is erased. When the message requesting battery replacement is displayed, replace the battery using the following procedure. ① Turn on the power. ON → ② Replace the existing lithium battery with a new one. → ③ Confirm the operation. - 52 - TSFE-X01010 LED list LED No. LD1 LD2 LD3〜6 Description Indicate the status in the 7-segment display area as shown below. Flickers while the X-ray tube is released. Meaning of indication by the LD1 and the LD2 in the 7-segment display area SS: An input is given from a sheet key. On: Exposure is being performed. OF: Exposure is finished. Check pin list CP No. CP1 CP2 CP3 CP4 CP5 DIP switch list SW No. 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 CP Name System clock System clock Vbatt +5V DGND Description CLK6M CLK12M Lithium battery voltage +5V GND Status Always ON Always OFF Always OFF Always OFF ON: Measured value display mode Always OFF Always OFF ON: Initial setting mode ON: EEPROM initialization mode. ON: Phototimer option is provided. ON: 200 mAs option is provided. ON: Remote control option is provided. ON: Printer option is provided. ON: Bucky unit is connected. Always OFF ON: Adjustment mode - 53 - OFF: Normal mode OFF: Normal mode OFF: Normal mode OFF: Not provided. OFF: Not provided. OFF: Not provided. OFF: Printer option is not provided. OFF: Bucky unit is not connected. OFF: Normal mode TSFE-X01010 11.2. X CONT-2000 PCB CP24〜CP31 VR1,VR2 VR3 CP3〜CP17 CP1,CP2 CP32〜CP40 LD2〜LD4 LD1 CP18〜CP23 Function This PCB equipped with the following circuits performs general control. 1. I/O circuit, A/D circuit and D/A circuit to transfer data between the NEXSC PCB 2. Circuits to detect and feed back the TKV and the TMA 3. TKV OVER detection circuit (discharge detection circuit) 4. System timer circuit (X-ray cutoff signal) 5. IGBT drive control circuit 6. Filament heating FET drive control circuit 7. Dynamic display control circuit, sheet key input circuit and sheet key search circuit 8. -15V detection circuit 9. Power ON reset circuit (master reset signal) 10. Charging control 11. Interface circuit for the Bucky unit During PCB replacement The VR1 (mAs), the VR2 (VFADJ) and the VR3 (PULSEADJ) should be adjusted. Refer to “Appendix C2 X CONT-2000 Adjustment Method” in the Installation Manual. - 54 - TSFE-X01010 LED list LED No. Signal Name LD1 KEY IN LD2 AR LD3 HX LD4 CHON-O Description Lights when an input is given from the panel. Lights when the READY switch or the first step of the hand switch is pressed. Lights when the X-RAY switch or the second step of the hand switch is pressed. Lights during charging. VR list VR No. VR1 VR2 VR3 Check pin list CP No. CP Name CP1 +5V CP2 GND CP3 CHECK1 CP4 CHECK2 CP5 VSIN CP6 CP7 IFST IFV CP8 VS CP9 TKV CP10 VN CP11 CP12 CP13 CP14 CP15 CP16 CP17 CP18 CP19 CP20 CP21 CP22 TMA PKV RMA FVR PH-REF GND +15V +5V GND VP1 VP2 IC CP23 LV CP24 CP25 +15V GND VR Name mAs VFADJ PULSE ADJ Description Adjusts the mAs clock. Adjusts the inverter working frequency. Adjusts the IGBT gate pulse. Description +5V GND To be short-circuited during adjustment. Shirt-circuit them first, then turn on the power. Short-circuiting them in the power ON status is invalid. Refer to Appendix C in the Installation Manual. 15V input during adjustment. Refer to Appendix C in the Installation Manual. Filament heating voltage signal (after correction) Measured filament current (1V/1A) Inverter working frequency voltage until the tube voltage reaches the preset value Measured tube voltage (1V/20 kV) Reference inverter working frequency voltage after the tube voltage reaches the preset value Measured tube current (1V/50 mA) Preset tube voltage (1V/20 kV) Preset tube current (1V/50 mA) Preset filament current (1V/1A) Preset phototimer reference voltage GND +15V +5V GND Charging voltage detection (1V/100V) Not used Not used Power supply confirmation (2V/100V) (which detects 100V output from the T1) +15V GND - 55 - TSFE-X01010 CP26 CP27 -15V VIN CP28 MASCLK CP29 F CP30 mA-IN CP31 CHGREF CP32 CP33 CP34 CP35 CP36 CP37 CP38 CP39 CP40 GND A/D B/C KVFBKVFB+ KVT Q1-4 EXRR Q2-3 -15V Inverter frequency voltage Signal created when the TMA is converted into the frequency by the V/F converter. This signal (5 kHz/50 mA) is used in the mAs timer. Oscillation frequency to drive the IGBT. This signal is adjusted by the VR2 (VFADJ). Refer to Appendix C in the Installation Manual. To be short-circuited during adjustment Charging control reference voltage (equivalent to 330V) which is approximately 3.3V (330V) GND IGBT drive control signal IGBT drive control signal Tube voltage feedback signal (-) Tube voltage feedback signal (+) Signal which rises when the tube voltage reaches the preset value Heating FET drive control signal Exposure signal Heating FET drive control signal - 56 - TSFE-X01010 11.3. INVERTER UNIT MUX Function: This PCB equipped with the following circuits controls the power system. 1. IGBT circuit 2. Filament heating circuit 3. Starter circuit During PCB replacement Adjust the FVR again. Set the DIP switches also. Check pin list CP No. CP Name Description CP1 CL Filament heating voltage CP2 CO CP3 Gc IGBT gate signal CP4 Ec CP5 Ga IGBT gate signal CP6 Ea CP7 YC Stator coil current detection (1V or more is normal.) CP8 Gd IGBT gate signal CP9 Ed CP10 Gb IGBT gate signal CP11 ZC Stator coil current detection (1V or more is normal.) CP12 Eb IGBT gate signal CP13 GND GND Fuse Fuse Name Description F1 Provided for protection DIPSW SW No. Function SW1A ON in the MUX-10 SW2A Not used - 57 - TSFE-X01010 11.4. INPUT PCH-99 Function This PCB equipped with the following circuits controls the power supply. 1. Collimator lamp voltage output circuit 2. Electromagnetic lock voltage output circuit 3. Charging voltage detection circuit 4. Supply voltage confirmation During PCB replacement The VR1 (VP1ADJ) should be adjusted. Refer to “Appendix C3 Charging Voltage Adjustment Method” in the Installation Manual. LED LED Name LED1 Description Lit while electric charge remains. Check this LED during a work. VR VR No. VR1 VR Name VP1 ADJ Description Adjusts the charging voltage detection function. - 58 - TSFE-X01010 Check pin list CP No. CP1 CP2 CP3 CP4 CP5 CP6 Fuse list Fuse No. F1、F2、F3 F4 F5 F6 F8 Signal Name VC0 VC +15V GND VPA VPB Description Charging voltage +15V GND Provided for charging voltage check. VP1=0.246×(VPA−VPB) Description Not used 280 VAC line from the transformer T1 13 VAC line from the transformer T1 125 VAC line from the transformer T1 100 VAC line from the transformer T1 11.5. SHEET PANEL This PCB is used also in the MUX-100. During PCB replacement Adjust the buzzer sound volume trimmer. - 59 - TSFE-X01010 11.6. MUX AEC I/F VR1 VR2 Function This PCB is provided for the phototimer option. This PCB compares the paddle signal from the detection unit with the reference voltage PH-REF from the NEXSC PCB, and outputs the exposure cutoff signal IPOF to the X CONT PCB. During PCB replacement Adjust the VR1 and the VR2 in accordance with the description on installation of the phototimer option in the Installation Manual. VR list VR No. VR Name Description VR1 OFFSET Adjusts the offset. VR2 GAIN Adjusts the gain. Check pin list CP No. Signal Name Description CP1 IN1 Phototimer signal 1 CP2 IN2 Phototimer signal 2 CP3 REF Phototimer reference voltage CP4 IPOF Phototimer cutoff signal CP5 GND GND - 60 - TSFE-X01010 11.7. RC RECEIVER Function This PCB functions as the receiving side of the remote control unit option. During PCB replacement There are no adjustment items. - 61 -