Uploaded by Paulo Skywalker

Manual serviço Mobileco Shimadzu

advertisement
TSFE-X01010
SHIMADZU MEDICAL SYSTEMS NISHINIHON CORPORATION
TECHNICAL SUPPORT & SERVICES DEPT.
TSFE-X01010
- ii -
TSFE-X01010
CONTENTS
1.
Introduction ......................................................................................................1
1.1.
1.2.
2.
3.
Outline .....................................................................................................................................1
Reference ...............................................................................................................................2
Cautions ............................................................................................................3
Features of MUX-10 .........................................................................................5
3.1.
3.2.
3.3.
3.4.
3.5.
3.6.
4.
Super Maintenance Function .................................................................................................5
FVR Adjustment......................................................................................................................5
Starter ......................................................................................................................................5
Measured Value Display Function..........................................................................................5
TMA Waveform .......................................................................................................................5
Exposure Counter/Error Log Display Function ......................................................................5
Periodical Inspection ........................................................................................6
4.1.
4.2.
4.3.
4.4.
5.
6.
7.
8.
Mechanical Inspection Items ..................................................................................................6
Electrical Inspection Items ......................................................................................................7
Inspection Items Related to X-ray Tube.................................................................................7
Cleaning ..................................................................................................................................7
Periodically Replaced Parts..............................................................................8
Recommended Spare Parts ..............................................................................9
PARTS MAP ................................................................................................... 11
REPLACING THE PARTS AND ADJUSTING............................................14
8.1. REPLACING THE BRAKE SHOE.......................................................................................14
8.2. Adjusting the brake ...............................................................................................................15
8.3. Replacing the caster..............................................................................................................16
8.4. Replacing the X-ray tube. .....................................................................................................17
8.4.1.
Replacing the collimator. ............................................................................................17
8.4.2.
Removing the X-ray tube............................................................................................18
8.4.3.
Reassemble X-ray tube unit.......................................................................................20
8.4.4.
Procedure after exchange x-ray tube.........................................................................23
8.4.5.
Adjusting after replacing the X-ray tube. .................................................................24
8.5. Adjusting of tube rotation. .....................................................................................................24
8.6. Adjusting of arm catcher .......................................................................................................25
8.7. Arm bearing adjusting. ..........................................................................................................25
8.8. Replacing the wire.................................................................................................................26
8.9. Replacement of Collimator Lamp.........................................................................................28
8.10.
Checking IGBT..................................................................................................................29
8.11.
Waveform Confirmation and Adjustment.........................................................................30
9.
Troubleshooting ..............................................................................................31
9.1.
9.2.
9.3.
When Power Does Not Turn On or Off ................................................................................31
When X-ray Tube Rotation Sound Is Abnormally Large .....................................................31
When Collimator Lamp Does Not Turn On or Off................................................................32
- iii -
TSFE-X01010
9.4. When Lock Cannot Be Released or Set..............................................................................33
9.5. UD125P-CH Self-diagnosis Function...................................................................................34
9.5.1.
Alarm message list......................................................................................................34
9.5.2. Error message list......................................................................................................34
9.6. Countermeasures against Error Messages .....................................................................35
9.6.1. H1 and H2 ...................................................................................................................35
9.6.2. H3..................................................................................................................................35
9.6.3. F01 and F02 ................................................................................................................35
9.6.4.
F03 ................................................................................................................................36
9.6.5. F11 (KV OVER)..........................................................................................................36
9.6.6. F12 (KV UNDER)......................................................................................................37
9.6.7. F13 (NO KV) ...............................................................................................................38
9.6.8. F14 (KV BRAKE).......................................................................................................39
9.6.9. F21 (MA OVER) and F22 (MA UNDER) ..............................................................40
9.6.10.
F23 (NO MA) ..........................................................................................................41
9.6.11.
F31 (IF OVER) and F32 (IF UNDER)...............................................................42
9.6.12.
F33 (NO IF).............................................................................................................43
9.6.13.
F41 ............................................................................................................................44
9.6.14.
F43 ............................................................................................................................45
9.6.15.
F44 and F45 ............................................................................................................46
9.6.16.
F46 ............................................................................................................................47
9.6.17.
F61 ............................................................................................................................48
9.6.18.
F62 ............................................................................................................................49
9.6.19.
F6E ...........................................................................................................................49
9.6.20.
F6F ...........................................................................................................................49
10.
11.
MUX-10 Block Diagram.................................................................................50
Explanation on Each PCB..............................................................................52
11.1.
11.2.
11.3.
11.4.
11.5.
11.6.
11.7.
NEXSC-MUX10................................................................................................................52
X CONT-2000 PCB.........................................................................................................54
INVERTER UNIT MUX...................................................................................................57
INPUT PCH-99 ................................................................................................................58
SHEET PANEL ................................................................................................................59
MUX AEC I/F ...................................................................................................................60
RC RECEIVER ................................................................................................................61
- iv -
TSFE-X01010
1. Introduction
1.1.
Outline
This service manual describes the manual type Mobile X-ray Unit MUX-10. Use this service
manual as the reference data of service works for inspection, repair, etc.
-1-
TSFE-X01010
1.2.
Reference
The following reference are related to this service manual. All of these reference are required in
service works. Make sure to use them together with this service manual. The information
described in the following reference is not described in this service manual in principle.
① MUX-10 Installation Manual
Document No.: M503-E308
This manual describes the information required in installation and maintenance as well as detailed
adjustment methods. When replacing PCBs, make sure to perform adjustment using the
described adjustment methods.
② MUX-10 Connection Diagram Book
The document No. is not given.
This manual shows circuit diagrams of the MUX-10.
③ MUX-10 Parts List
Document No.: M503-4003
This list shows the information required in ordering parts.
④ MUX-10 Operation Manual
Document No.: M503-E006
This manual describes the specifications, the operating procedures and the information on
maintenance and inspection, guarantee, disposal, etc.
-2-
TSFE-X01010
2. Cautions
Make sure to read thoroughly the cautions below before starting a work during installation,
inspection or maintenance.
Fixing the Counterweight
Make sure to fix the counterweight inside the column to the column with bolt before removing the
X-ray tube and the collimator. If the counterweight is not fixed, the MUX-10 may become
unbalanced and extremely dangerous.
Caution on Waveform Adjustment
Make sure to take the following action to prevent charging before observing the IGBT gate
waveform. If the IGBT gate waveform is observed in two positions at the same time in the
charged status, an accident may be caused.
Remove the fuse F4 from the INPUT PCH-99 PCB.
Make short-circuit the CP3 (CHECK1) and the CP4 (CHECK2) on the X CONT PCB.
Caution on Power ON
Do not turn on the power while the extension NEXSC PCB is provided.
If such a PCB is provided, the MUX-10 may malfunction.
Caution on High Voltage Generation
Do not generate high voltage while the high voltage cable and the X-ray tube are not connected.
-3-
TSFE-X01010
Prevention of Electrical Shock (1)
Turn OFF the power before starting maintenance of the inside of the cabinet. Even if the power is
turned OFF, electric charge may remain for a while. Make sure that the LD1 on the INPUT
PCH-99 PCB is lit off before starting maintenance.
Prevention of Electrical Shock (2)
After disconnecting the high voltage cable, make sure to let the high voltage plug pin be in contact
with the grounded portion so that the plug pin is discharged.
Prevention of Charge
When touching a PCB, make sure to wear the wrist band for prevention of static electricity so that
charge can be prevented.
Protection against Radiation
When checking the filament of the X-ray tube or visually checking discharge inside the X-ray tube,
make sure to take proper protective action such as use of lead glass.
-4-
TSFE-X01010
3. Features of MUX-10
The features of the MUX-10 regarded as important to service are described below.
3.1.
Super Maintenance Function
The initial setting mode and the adjustment mode are available without setting the DIP switches on
the NEXSC PCB. Refer to Appendix D in the Installation Manual.
3.2.
FVR Adjustment
The MUX-10 offers two adjustment modes, the “two-point adjustment mode” in which the CPU
calculates other data after the operator has adjusted only two points in the FVR table and the
conventional “all-point adjustment mode” in which the operator adjusts all data in the FVR table.
Perform adjustment in the two-point adjustment mode in installation and adjustment, then confirm
the FVR in the all-point adjustment mode.
3.3.
Starter
The starter circuit is assembled in the INVERTER UNIT PCB. The starter starts in 1.6 sec at
125V, and is driven intermittently at every 10 sec in the steady state.
3.4.
Measured Value Display Function
Exposure for 32 msec or more is required to use the measured value display function. Have this
fact in mind in adjusting the FVR.
3.5.
TMA Waveform
The TMA waveform is of 1 V/50 mA. In conventional units, the preset TMA value was
displayed for 10 msec after exposure was started. In the MUX-10, the preset value is not
displayed.
3.6.
Exposure Counter/Error Log Display Function
The MUX-10 has the function to count the number of times of exposure and to restore the error log.
Refer to Appendix B in the Installation Manual.
-5-
TSFE-X01010
4. Periodical Inspection
Perform the periodical inspection in accordance with the tables below.
4.1.
Mechanical Inspection Items
Inspection item
1
2
3
4
5
6
7
Are bolts tightened securely in the tube holder?
Are flat countersunk head screws in the
collimator mounting area and flush head bolts in
the X-ray tube and the mounting ring tightened Tighten the bolts.
Apply LOCKTITE.
securely? (Remove the collimator for check.)
Is the column fixed securely?
Are the fixing bolts tightened securely?
Are front wheel casters fixed securely?
Does the X-ray tube rotate smoothly without
abnormal sound?
Is the tension of two wire ropes equivalent?
Are the double-nuts tightened securely?
Are the wire ropes free from rust?
Are the wire ropes free from flaws? Are
element wires free from cuts and fine splits?
Does the X-ray tube move up and down smoothly
9
without abnormal sound?
Does the column rotate smoothly without
10
abnormal sound?
8
11
Countermeasures
Is the X-ray tube expanded smoothly without
abnormal sound?
12 Does the lock release switch operate normally?
Apply the lubricant.
Adjust the wire nuts.
Apply the lubricant.
Replace the wire ropes.
Clean the rails. Apply the
lubricant. Or release the lock,
and check for defects.
Remove the cover from the arm
end face, and check the
movement visually.
Investigate the cause in
accordance with connection
diagrams.
13 Is the grip fixed securely?
Tighten the bolts.
14 Is the travel brake released and locked normally?
Adjust the gap. Clean the
wheels.
15 Does the collimator lamp light normally?
Replace the lamp.
16 Are cables not caught?
Treat the cables again.
17 Does the arm catch unit operate normally?
Check the movement visually.
-6-
Inspection result
TSFE-X01010
4.2.
Electrical Inspection Items
Inspection item
1
2
3
4
5
6
7
8
4.3.
Countermeasures
Measure the supply voltage at each check pin.
Observe actually the TKV and TMA waveforms
to check whether the tube voltage, the tube
current, the exposure time and the tube current
time product are output in accordance with the
settings.
Confirm the photograph density if the phototimer
is provided.
Adjust the DC power output.
(Adjustment is available only
for 24 VDC.)
Adjust again the FVR and
IGBT gate waveforms.
Adjust the phototimer again.
Is the high voltage cable plug tightened securely?
Tighten it securely.
Are terminals and connectors fixed or contacted
securely?
Do the panel display and the sheet keys function
normally?
Are cables free from breakage? Check for
breakage in the cable reel plug especially.
Fix the terminals and the
connectors.
Check contact of the connectors.
Or replace the panel.
Is the ground belt free from damages?
Replace the ground belt.
Replace the cables.
Inspection Items Related to X-ray Tube
Inspection item
Countermeasures
1
Is the X-ray tube rotation sound normal?
Replace the X-ray tube.
2
Is the X-ray tube free from oil leak?
Replace the X-ray tube.
4.4.
Inspection result
Inspection result
Cleaning
Make sure to clean the MUX-10 in the periodical inspection.
Cleaning position
1
Appearance of MUX-10
2
Inside of MUX-10
3
Wheels, brake face and
electromagnetic face of lock and rail.
Caution
Remove dirt with cloth soaked in neutral
detergent .And use carefully so that the
MUX-10 is not damaged.
Remove dusts by using a vacuum cleaner.
Fire may be caused by dusts and foreign
substances.
Remove dirt.
-7-
Result
TSFE-X01010
5. Periodically Replaced Parts
Replace the following parts in the periodical inspection.
Part Name
Part No.
Used Area
Q’ty
Halogen lamp
062-65003-02
Collimator
1
LITHIUM BATTERY ASSY
501-78646
NEXSC
1
FUSE, SC-10
072-01649-01
FUSE, 326010
072-01665-34
FUSE, 313002
072-01664-26
SILICONE RUBBER JACKET
SILICONE WASHER
511-15061-13
X-RAY TUBE
2
SILICONE PACKING
532-23029
HIGH VOLTAGE
TRANSFORMER
2
1
INPUT PCH-99
2
2
If the MUX-10 has been used for 10 years or more, replace the following parts.
Part Name
Part No.
Product type
Q’ty
503-46520-01
MUX-10
2
503-46520-04
MUX-10J
2
Wire rope
-8-
TSFE-X01010
6. Recommended Spare Parts
The following parts are recommended as spare ones (including periodically replaced parts).
RANK A
QTY
P/N
DESCRIPTION
503-49045-01
062-65003-02
064-54801-98
064-54803-96
501-78646
072-06033-08
072-01664-20
072-02004-16
072-01649-01
072-01665-34
072-01664-26
074-80678-16
074-80678-86
502-21470
502-21572
502-21859
502-20775
060-39940-01
511-15061-13
CODE REEL ASSY
LAMP HALOGEN #55939
SW PB A3AT-90A1-00L
FLANGE A3A-222
LITHIUM BATTERY ASSY
FUSE , 600FH-100
FUSE , 313001
FUSE , 218001
FUSE , SC-10
FUSE , 326010
FUSE , 313002
POWER SUPPLY , LDC60F2
POWER SUPPLY , LDA-50F-24-Y
INVERTER UNIT MUX ASSY
X CONT-2000 ASSY
NEXSC-MUX10
INPUT PCH-99 ASSY
IGBT , 2MBI300N-060
MAINTENANCE KIT MINI
MUX-10
1
1
3
3
1
1
1
2
1
2
2
1
1
1
1
1
1
2
1
MUX-10J
1
1
3
3
1
1
1
2
1
2
2
1
1
1
1
1
1
2
1
RANK B
P/N
DESCRIPTION
503-34889-02
510-20036-05
503-51143
503-50495
503-51287
503-50093
511-64073
EARTH BELT
PLUNGER PBF-12-KSN
BRAKE SHOE
MICRO SWITCH ASSY B
MICRO SWITCH ASSY C
RUBER CUSHION
SHEET PANEL MUX-A
QTY
MUX-10
1
2
2
1
1
1
1
-9-
MUX-10J
1
2
2
1
1
1
1
TSFE-X01010
RANK C
QTY
P/N
DESCRIPTION
503-46520-01
503-46520-04
511-18026-01
511-37050
510-37081-04
502-20700-01
502-21479-01
065-90756-01
065-68009-48
065-40261-01
WIRE ROPE
WIRE ROPE
CABLE COVER
WHEEL , 380
CASTER , 420SA
H.T. TRANS D125PH-C1
H.T. CABLE MUX
CIRCUIT PROTECTOR , CP22HS/15DB1
RELAY , G3NA-420B-DC5V
CONTACTOR , SH5AC100V1A5B
- 10 -
MUX-10
2
1
2
2
1
1
1
1
1
MUX-10J
2
1
2
2
1
1
1
1
1
TSFE-X01010
7. PARTS MAP
MAIN BODY
RESISTANCE
R1A,R1B,R2A,R2B
SSR
NEXSC
DCPOWER
SUPPLY PS1
DC POWER
SUPPLY PS2
X CONT 2000
ARM SW
SW4
INPUT
PCH99
INVERTER
UNIT
T1TRNS TERMINAL
CORDREEL
- 11 -
TSFE-X01010
CAPACITOR
RELAY
MG-D
RESISTANCE
RDA,RDB,RDC
- 12 -
TSFE-X01010
ELECTRIC
LOCK
TRANS T1
FOR REMOVING
THE ELECTRIC
LOCK
EARTH BELT
ELECTRIC
LOCK
- 13 -
TSFE-X01010
8. REPLACING THE PARTS AND ADJUSTING
8.1.
REPLACING THE BRAKE SHOE
SPRING PLATE
BRAKE COVER
BRAKE BAR
1. Remove the brake cover. Remove the cassette box.
2. Remove the spring plate. Retighten the both bolts equally.
3. Remove the spring and the bolts.
4. Pull out the brake bar.
5. Remove the brake shoe assy from the brake bar.
BRAKE SHOE
SPRIG
NUT
BRAKE SHOE
CASE
SPRING PLATE
- 14 -
BRAKE BAR
TSFE-X01010
8.2.
Adjusting the brake
Adjust the brake in case that brake release is not easy.
Retighten the nut and rotate the eccentric shaft for changing the gap between the wheel
and the brake shoe. After adjusting , tighten the nut securely.
NUT
ECCENTRIC
SHAFT
Apply the greese to below.
grease
- 15 -
TSFE-X01010
8.3.
Replacing the caster.
The special tool is need to replace the caster.
Spanner :503-37353
At first retighten the screw. Fix the shaft by special spanner and retighten the nut.
remove the nut
fix this shaft by
spanner
screw
- 16 -
TSFE-X01010
8.4.
Replacing the X-ray tube.
It is very dangerous to remove the X-ray tube and the collimator because the
balance can not be keep. Fix the weight in the column by the bolt.
8.4.1.
Replacing the collimator.
Remove the collimator from the X-ray tube.
Remove
screws
these
4
Remove the socket and high voltage cable from X-ray tube and remove the low voltage
cable at the junction connector.
Remove these screws
- 17 -
TSFE-X01010
8.4.2.
Removing the X-ray tube.
Remove the cable fixing at the arm and the column.
Remove the U-nut at the end of the arm.
remove the
cover,U-nut,dics
spring,detent
washer,thrust
washer
Remove the X-ray tube assy.
thrust washer
detent washer
U-nut
disc spring
- 18 -
TSFE-X01010
X-ray tube
assy
U-nut,disc spring,detent
washer,thrust washer
- 19 -
TSFE-X01010
8.4.3.
Reassemble X-ray tube unit
Remove the cover from the tube rotation unit attaching the tube to the arm, remove socket head cap
screws (in 4 positions on each of the left and right sides) inside, then remove and move down the tube
from the holding frame.
CAUTION
The weight of the tube ASSY is approximately 13 kg. Pay rigid
attention in moving it down from the arm.
Remove these socket
head cap screws (in 4
positions).
Remove these socket
head cap screws (in 4
positions).
Rotation unit on
the right side
Rotation unit
on the left side
Remove the left and right tube rotation units from the tube.
As to the right side, loosen the U nut (new type lock nut not requiring a inner clip washer) in the
center, then remove the rotation stopper, the spring washer, etc. After that, remove socket head cap
screws from the shaft area, then remove the shaft area.
As to the left side, remove the cover (countersunk head screws in 2 positions), then remove socket
head cap screws from the inside.
CAUTION
Grease will be applied on sliding units at the end. Every time removing
each part, clean the sliding units. Moreover, please be sure to decompose
at the time of decomposition work, performing marking so that an
attachment position and a direction may be known from back. It
mistakes, and when assembling, it may not operate normally.
Left rotation unit
Remove these socket
head cap screws (in
4 positions).
Parts of left rotation
unit
- 20 -
TSFE-X01010
Right rotation unit
Before removing this
rotation stopper, mark
the
attachment
position.
Remove
this U nut.
It is double-faced.
Remove these socket
head cap screws (in 4
positions).
Parts of
right
rotation
unit
- 21 -
TSFE-X01010
Remove the collimator rotation unit,
and attach it to a new tube. Please
be careful in the direction of stopper
at this time.
Remove these hexagon socket
countersunk head screws (in 4
positions).
Collimator
rotation unit
Attach the left and right rotation units to the new tube in the order of removal. At this time, clean the
sliding units, and apply grease on the following portions before attachment.
CAUTION
Make sure to use grease not containing abrasive.
Recommended grease: Albania grease 2 (Showa Shell)
Apply grease on this
portion.
Apply grease on
this portion.
Apply grease
on the entire
this face.
Apply grease
on the entire
this face.
- 22 -
Apply grease
on
this
portion.
TSFE-X01010
8.4.4.
Procedure after exchange x-ray tube.
At the end, attach the low voltage cable, the collimator and the high voltage plug, and treat the cables.
At this time, attach the high voltage plug using the following procedure.
ⅰ Make sure that the inside of the socket in the X-ray tube unit is not dirty.
ⅱ Make sure that the plug and the ring nut are not dirty.
(For the name of each part, refer to Fig. 6-31.)
ⅲ Let the ring flange go through the ring nut.
ⅳ Apply silicon oil on the silicon rubber jacket and the silicon washer.
ⅴ Fit the silicon washer into the tip of the plug.
ⅵ Put the plug into the socket. When the plug tip reaches the bottom of the socket, turn the plug to
find the position in which the plug is fitted into the key way, then push the plug.
ⅶ Tighten the ring nut with hand.
ⅷ Tighten the rotation stopper screw so that the ring nut does not rotate.
Ring nut
Silicon
washer
Ring flange
Silicon rubber
jacket
Not apply
toa
pin
head.
- 23 -
High voltage
cable
Apply oil on
this
silicon
rubber.
Silicon oil
TSFE-X01010
8.4.5.
Adjusting after replacing the X-ray tube.
Perform the aging.
Perform FVR adjusting.
8.5.
Adjusting of tube rotation.
Forth for rotate the tube can be adjusted by U-nut.
In case that rotation of X-ray tube is not smooth, please apply the greese refer to
installation manual appendix F.
- 24 -
TSFE-X01010
8.6.
Adjusting of arm catcher
In arm catcher, there are two swiches.
SW4
ARM SW
SW4 should be ON when arm is hold. ARM switch should be ON when the lever is
pulled.
8.7.
Arm bearing adjusting.
Adjust the eccentric bearing to decrease the gap of arm sliding.
eccentric bearing
- 25 -
TSFE-X01010
8.8.
Replacing the wire
For replacing the wire, it need to fell down the column. Please perform the procedure
referring to below. And take care to safty.
Fix the weight in the column by the bolt.
Remove the X-ray tube referring to “replacing the X-ray tube”.
Remove the bolts at foot side of the column.
fix the weight by the
bolt
Retighten the screw and remove the washer and pin. Pull out the shaft. Do not use the
metal hammer, please use rubber hammer to protect the shaft from changing the shape.
If shape of the end of the shaft is changed, it is impossible to assemble again.
bearing with pully
washer
pin
shaft
pully
collar
- 26 -
screw
TSFE-X01010
Remove the bolt fixing the weight. Pull out the weight from the column.
weight
After remove the subweight and bearing mount, retighten and remove the metal fitting
of wire. Remove the wire from the carrige.
double nuts
metal fitting
In case of reassembling, do not forget to fix the weight by the bolt.
- 27 -
TSFE-X01010
8.9.
Replacement of Collimator Lamp
Replace the collimator lamp using the following procedure.
CAUTION :Never touch directly the collimator lamp. If dirt is adhered on the lamp,
the illuminance may be deteriorated.
NOTE : After the power of the system has been OFF for a while or after a while after
the key switch was set to OFF (that is, after the collimator lamp is cooled down), replace
the collimator lamp.
1.
2.
3.
4.
Remove the lamp outer cover of the collimator.
Remove the lamp cover.
Replace the lamp (P/N 062-65003-02 halogen lamp).
Attach the covers again.
Lamp outer cover
(Remove screws
in 3 positions.)
Lamp
cover
(Remove screws in 2
positions.)
- 28 -
Replace the collimator
lamp. (Never touch it
directly.)
TSFE-X01010
8.10. Checking IGBT
In case that wave form of IGBT gate signal is abnormal, it means defective of IGBT.
If you want to check IGBT only, refer to below data. This IGBT unit has two IGBT in one
unit. Please refer to circuit diagram below.
Reference value ( measured at prototype MUX-10)
C2E1
+
Digital meter
−
C2E1
+
−
Digital meter
C1
terminal
C1
−
+
Measured value
0.8MΩ
infinity
E2
−
+
抵抗値
infinity
0.8MΩ
E2
C2E1
IGBT
circuit
diagram
- 29 -
TSFE-X01010
8.11. Waveform Confirmation and Adjustment
Refer to Appendix C in the MUX-10 Installation Manual. Make sure to take proper action
such as short-circuiting in accordance with the instructions described there.
- 30 -
TSFE-X01010
9. Troubleshooting
Read this chapter as reference of troubleshooting.
9.1.
When Power Does Not Turn On or Off
The outline of the power line is shown below. Check the line from the check pins on the PCB
from power source and find the cause of the trouble. The main switch may be defective, the fuse
may be blown out, or the DC power supply may be defective.
9.2.
When X-ray Tube Rotation Sound Is Abnormally Large
Prepare the noise meter, and measure the rotation sound. When the measured value is 65 dB or
more at the distance of 1 m, the X-ray tube is abnormal.
→ Contact the Quality Assurance Department in Medical Systems Div. of Shimadzu
Corporation.
The rotation sound may be temporarily large in the early period (within 3 months after delivery),
and may return to the normal sound after a while (due to sticking of the solid lubricant of the
bearing). The skid time can be used to judge the rotation status of the X-ray tube. If the period
of time from rotation at low speed to stop by inertia is 1 min or more, the X-ray tube is normal.
- 31 -
TSFE-X01010
9.3.
When Collimator Lamp Does Not Turn On or Off
The outline of the wiring related to the lamp is shown below. Check the wiring from the lamp
terminals inside the collimator to power source, and find the cause of the trouble. The switch
may be defective, or the fuse may be blown out, or the PCB may be defective,or the lump may be
defective.
T1-MUX-10
COLLIMATOR
AC13V
CN1
CN5
J135
JMU
JT1IN
F5
LAMP
K6
INPUT PCH-99
SWITCH
JINP
JINP
X CONT-2000
- 32 -
NEXSC
CIRCUIT
PROTECTOR
TSFE-X01010
9.4.
When Lock Cannot Be Released or Set
The outline of the wiring related to the locks is shown below. Check the wiring from the
electromagnetic locks to DC power supply, and find the cause of the trouble. A switch may be
defective, an electromagnetic lock may be defective, or the PCB may be defective.
SL1
SL2
SL3
SL4
ARM
SWITCH
ARM
SWITCH
CN2
CN3
CN17
CN18
CN1
CN3
JMU
DC24V
K5A
INPUT PCH-99
- 33 -
TSFE-X01010
9.5.
UD125P-CH Self-diagnosis Function
9.5.1.
Alarm message list
Display
Description
With the set exposure condition, the heat accumulated on the anode of the X-ray tube is
over tolerance. Reduce the exposure condition, or stop exposure until warning code
disappears.
The heat accumulated on the anode of the X-ray tube has reached its tolerance limit. Stop
H2
exposures until this warning code disappears.
The temperature of X-ray tube assembly is too high. Stop exposures until this warning
H4
code disappears.
The battery for CPU memory backup needs replacing now. Contact your nearest
E
authorized Shimadzu service center.
The life of battery for CPU memory backup has expired. Contact your nearest authorized
EEE
Shimadzu service center.
NOTE) The CPU memory battery (lithium battery) backs up the SRAM on the NEXSC PCB.
When the life of this lithium battery is expired, the contents displayed just before the power was
turned off last time, the error log and the time setting are erased. Accordingly, the MUX-10 can
be used though there is inconvenience.
H1
9.5.2.
Error message list
For the action to be taken, refer to “7.16 counter measure against error message”.
Display
Description
F01,F02
The supply voltage is out of its normal range.
F03
The power supply circuit in the system is at fault.
F11〜F14
The X-ray tube voltage control is at fault.
F21〜F23
The X-ray tube current control is at fault.
F31〜F33
The filament heating control is at foult.
F41〜F46
The charge control is at fault.
F61
The starter is at fault.
F62
There is abnormality in the Bucky unit operation.
F6E、F6F
The irradiation control or photo timer
- 34 -
TSFE-X01010
9.6.
Countermeasures against Error Messages
Error messages displayed on the panel are described in detail, and the countermeasures are
described below.
9.6.1.
H1 and H2
The CPU calculates the accumulated heat capacity based on previous exposure, then issues these
alarm messages. If an either of these messages is issued even while there is not so many exposure,
check the voltage of lithium battery and initialize the CPU.
9.6.2.
H3
When the thermal switch of the X-ray tube is activated, this message is issued. If this message is
issued even while the X-ray tube is not hot at all, check the thermal switch signal.
9.6.3.
F01 and F02
(detected by software)
F01: The supply voltage is higher by approximately 20% or more than the normal value .
F02: The supply voltage is lower by approximately 30% or more than the normal value .
Check CP23 (LV) on X
CON T -2000.
W ithin range from
2.4 to 1.4V ?
YES
X CO NT or NEX SC is
abnormal.
→ Replace them in turn.
NO
Fuse F5 on
IN PU T PCH
normal?
NO
Replace F5.
YES
Voltage
between pins 3 and
5 of connector J135 from
transformer T1-M UX10
13 V AC?
YES
NO
T 1-M UX10 is defective.
Or check power line to T 1M UX 10.
- 35 -
IN PU T PCH is abnormal.
→ Replace it.
TSFE-X01010
9.6.4.
F03
(detected by hardware)
The -15V power supply for the control circuit is abnormal.
Check -15V (CP26)
on X CONT.
Normal
X CONT or NEXSC is
abnormal.
→ Replace them in turn.
Abnormal
Check output
of power supply
PS2.
Abnormal
PS2 is abnormal.
→ Replace it.
Normal
-15V may be shortcircuited with GND on X
CONT or other PCB.
9.6.5.
F11 (KV OVER)
(detected by software, at 20ms after exposure and each 100ms )
The measured tube voltage is higher by approximately 30 kV or more than the preset value.
Measure TKV(CP9) and
RKV(CP12) on X CONT.
RKV
abnormal? Or both
TKV and RKV
normal?
YES
Only TkV is abnormal.
① X CONT,
② INVERTER UNIT
③ High voltage transformer
may be defective.
- 36 -
X CONT or NEXSC is
abnormal.
→ Replace them in turn.
TSFE-X01010
9.6.6.
F12 (KV UNDER)
The measured tube voltage is lower by 50% or more than the preset value (detected by
software, at 20ms after exposure and each 100ms )
NOTE) It is most possible that glow discharge has occurred in the X-ray tube.
When checking the IGBT gate waveform, take proper action such as short-circuiting to prevent
generation of high voltage in accordance with Appendix C in the Installation Manual.
Check TKV(CP9) and
RKV(CP12) on X
CONT.
YES
RKV abnormal?
Or both normal?
X CONT or NEXSC is
abnormal.
→ Replace them in turn.
NO
Only either
KVFB+ or KVFBoutput?
YES
One side of high voltage transformer is defective.
→Replace high voltage transformer.
NO
FVR is too large. Adjust it.
If it does not become normal, proceed to next step.
Perform adjustment
related to tube voltage
shown in Appendix C in
Installation Manual.
If it does not become normal, proceed to next step.
Abnormal
Waveform
at F(CP29), A/D(CP33)
or B/C(CP34) normal?
X CONT is abnormal.
→ Replace it.
Normal
IGBT
gate waveform on
INVERTER UNIT
normal?
Abnormal
Normal
IGBT
single element
check
Normal
Abnormal
① X ray tube
② High voltage transformer
③ High voltage cable
may be defective.
Replace IGBT.
- 37 -
INVERTER UNIT is
abnormal.
→ Replace it.
TSFE-X01010
9.6.7.
F13 (NO KV)
The measured tube voltage is approximately 10 kV or less (detected by software, at 20ms after
exposure and each 100ms )
When checking the IGBT gate waveform, take proper action such as short-circuiting to prevent
generation of high voltage in accordance with Appendix C in the Installation Manual.
Fuse on
INVERTER UNIT
blown out?
Blown out
Normal
Damaged
IGBT damaged?
Replace INVERTER
UNIT. Or replace only
IGBT and fuse.
Check TKV(CP9) and
RKV(CP12) on X CONT.
Normal
Replace fuse.
RKV abnormal?
Or both normal?
YES
X CONT or NEXSC is abnormal.
→ Replace them in turn.
Only TKV is abnormal.
Perform adjustment
related to tube voltage
shown in Appendix C in
Installation Manual.
Waveform
Abnormal
of F( CP29),A/D(CP33) or
B/C(CP34)
normal?
X CONT is abnormal.
→ Replace it.
Normal
IGBT gate
waveform on INVERTER
UNIT normal?
Abnormal
Normal
Normal
IGBT
single element check
Abnormal
High voltage transformer
may be defective.
Replace IGBT.
- 38 -
INVERTER UNIT is
abnormal.
→ Replace it.
TSFE-X01010
9.6.8.
F14 (KV BRAKE)
Abnormal measured tube voltage or X-ray tube discharge is detected (by the hardware).
This error message actually postulates discharge in the high voltage system (especially the X-ray
tube). This error may be eliminated by aging. Perform aging first.
Perform aging.
Same error occurs again.
High voltage cable may be poorly
contacted. Refer to "Appendix G
High Voltage Cable Connection
Procedure" in Installation M anual,
and connect it again.
Same error occurs again.
① X-ray tube,
② High voltage cable,
③ High voltage
transformer,
④ X CONT or NEXSC
may be defective.
Judgement in the last item in the flow chart above is difficult. Abnormality in a PCB means that a
false error message is issued, but the possibility is regarded as low.
If an either of the following phenomena is detected, it can be clearly determined that discharge has
occurred in the high voltage system.
1. Discharge sound “pop” is issued during exposure.
2. The error message is not issued at low kV, but is issued while the kV is rising.
3. The glass is clearly clouded (due to evaporation of the target material) when seen from the
X-ray tube irradiation port.
4. The target is clearly rough when seen from the X-ray tube irradiation port.
5. Spark discharge like thick thunderbolt can be seen at the time of exposure from the X-ray tube
irradiation port.
(Make sure to take protective measures against radiation such as use of lead glass.)
An aluminum filter is provided in the X-ray tube irradiation port. When checking the inside
of the glass tube, remove the lead cone fast, then remove the aluminum filter also.
- 39 -
TSFE-X01010
9.6.9.
F21 (MA OVER) and F22 (MA UNDER)
(detected by software, at 20ms after exposure and each 100ms )
F21: The measured tube current is higher by approximately 50 mA or more than the preset value
F22: The measured tube current is approximately 50% or less of the preset value .
When these messages are displayed, check the heating system also in accordance with the flow
chart for F31 and F32.
Check RMA(CP13) and
TMA(CP11) on X CONT.
YES
RMA abnormal?
Or both normal?
X CONT or NEXSC is abnormal.
→ Replace them in turn.
Only TMA is abnormal.
Adjust FVR again.
Same error occurs again.
X-ray tube may be
defective.
- 40 -
TSFE-X01010
9.6.10.
F23 (NO MA)
The measured tube current is approximately 10 mA or less (detected by software, at 20ms
after exposure and each 100ms )
Short-circuit bar
(mA+ and mA-) attached to
terminal block X1
on X CONT?
NO
Attach short-circuit bar.
YES
Check RMA(CP13) and
TMA(CP11) on X CONT.
RMA abnormal?
Or both normal?
YES
X CONT or NEXSC PCB is
abnormal.
→ Replace them in turn.
Only TMA is abnormal.
High voltage cable
conductive?
NO
Replace high voltage cable.
YES
Breakage or poor
contact in cable JFB between
high voltage transformer
and X CONT?
YES
Replace cable.
NO
① High voltage transformer may
be defective or
② X-ray tube filament may be
broken.
X-ray tube filament check method
Remove the collimator. Look into the X-ray tube from the X-ray tube irradiation port, and check
whether the filament is lit in the ready status. Make sure to use lead glass, etc. for protection
against radiation.
- 41 -
TSFE-X01010
9.6.11.
F31 (IF OVER) and F32 (IF UNDER)
(detected by the software at each 800ms when AR! Income, and at each 400ms when ready up)
F31: The measured filament current is higher by 50% or more than the preset value .
F32: The measured filament current is lower by 50% or more than the preset value .
Check FVR(CP15) and
IFV(CP7) on X CONT.
Both
normal? Or only FVR
abnormal?
YES
X CONT or NEXSC is
abnormal.
→ Replace it.
Only IFV is abnormal.
NO
C0 and CL
waveforms normal?
INVERTER UNIT or X
CONT is abnormal.
→ Replace it.
YES
① X CONT,
② INVERTER UNIT or
③ High voltage (filament)
transformer
may be defective.
In case of F31, ②
and ③ rarely occur.
In the last item, ① and ② means defect in the detection system. Actually, the possibility of ③
is high.
- 42 -
TSFE-X01010
9.6.12.
F33 (NO IF)
The measured filament current is approximately 1 A or less .
(detected by the software at each 800ms when AR! Income, and at each 400ms when ready up)
YES
Fuse F7
on INPUT PCH blown
out?
Replace F7.
NO
NO
C0 and CL
waveforms normal?
Heating FET may be
damaged. Refer to
Appendix C in Installation
Manual, and check CL and
C0 waveforms.
INVERTER UNIT or X
CONT is abnormal.
→ Replace it.
Waveform is abnormal.
YES
Replace INVERTER
UNIT.
Normal
Check filament
current of IFV(CP7) on
X CONT.
X CONT or NEXSC is
abnormal.
→ Replace it.
Abnormal
High voltage cable may be poorly contacted.
Refer to "Appendix G High Voltage Cable
Connection Procedure" in Installation
Manual, and connect it again. Exchange +
and - sides of high voltage cable.
Same error occurs again.
Check X-ray
tube filament
visually.
Filament is
broken.
Replace X-ray tube.
Normal
High voltage (filament)
transformer may be
defective.
- 43 -
TSFE-X01010
9.6.13.
F41
The measured capacitor voltage is approximately 360V or more. The charging capacitor is
charged beyond its withstand voltage.
NO
Transformer
tap changed correctly for
supply voltage?
Change tap in accordance
with Installation M anual.
YES
Perform "Appendix C
Charging Voltage
Adjustment" in Installation
M anual.
NO
Adjustment
performed correctly?
INPUT PCH-99 is abnormal.
→ Replace it.
Error message is displayed
even after adjustment.
NO
SSR normal?
Replace SSR.
YES
① INPUT PCH-99,
② X CONT or
③ NEXSC
may be defective.
→ Replace them in turn.
- 44 -
TSFE-X01010
9.6.14.
F43
The measured capacitor voltage is approximately 20V or less at 3 sec after start of charging.
The power may not be supplied.
Transformer
tap changed correctly for
supply voltage?
NO
Change tap in accordance
with Installation M anual.
YES
Fuse on
INPUT PCH blown
out?
YES
Replace fuse.
NO
Perform "Appendix C
Charging Voltage
Adjustment" in Installation
M anual.
NO
Adjustment
performed correctly?
INPUT PCH-99 is abnormal.
→ Replace it.
Error message is displayed
even after adjustment.
NO
SSR normal?
Replace SSR.
① INPUT PCH-99,
② X CON T or
③ NEXSC
may be defective.
→ Replace them in turn.
- 45 -
TSFE-X01010
9.6.15.
F44 and F45
F44: The measured capacitor voltage does not reach approximately 240 V (at which the charging
control resistor is reset).
F45: The measured capacitor voltage does not reach 310 V (which is the preset value).
Transformer
tap changed correctly for
supply voltage?
NO
Change tap in accordance
with Installation M anual.
YES
Perform "Appendix C
Charging Voltage
Adjustment" in Installation
M anual.
NO
Adjustment
performed correctly?
INPUT PCH-99 is abnormal.
→ Replace it.
Error message is displayed
even after adjustment.
① INPUT PCH-99,
② X CONT or
③ NEXSC
may be defective.
→ Replace them in turn.
- 46 -
TSFE-X01010
9.6.16.
F46
The capacitor voltage rises while charging is stopped (detected by the software).
The SSR may be short-circuited.
NO
SSR is abnormal.
→ Replace it.
SSR normal?
YES
X CON T or NEXSC is
abnormal.
→ Replace it.
- 47 -
TSFE-X01010
9.6.17.
F61
Starter error (The starter does not issue the current detection confirmation signal.)
The current does not flow in the stator coil, and the rotor does not rotate.
NO
SW1A set to ON on
INVERTER UNIT?
Set SW1A ON.
YES
F6 blown
out on INPUT PCH?
Voltage between CP7
and CP13,CP11 and
CP13 should be 1V or
more
YES
Replace F6.
NO
Check coil
current of YC(CP7) and
ZC(CP11) on INVERTER
UNIT.
Normal
Replace X CONT or NEXSC.
Abnormal
X-ray tube coil
resistance
X-Y: 13±0.7W
X-Z: 46±2.3 W
X-ray tube connector
pin layout
Pin No. 1 = X
Pin No. 2 = Y
Pin No. 3 = Z
Measure
X-ray tube coil resistance.
Abnormal
Replace X-ray tube.
- 48 -
Normal
Low voltage cable is poorly
contacted. → Replace it.
or
INVERTER UNIT is abnormal.
→ Replace it.
TSFE-X01010
9.6.18.
F62
In 1 sec after HX1 is input, XOK is not sent back. This message is not displayed usually.
Confirm the setting of the DIP switch.
9.6.19.
F6E
Exposure is not finished even after the backup time or beyond the preset mAs.
The X CONT-2000 PCB (X-ray cutoff circuit) may be abnormal.
9.6.20.
F6F
The exposure control circuit is abnormal.
In 250 msec after the EXR is output, the KVT (kV rising signal) is not sent back. This
abnormality occurs when the shutdown signal IPOF is input at the same time with the EXR due to
input of direct radiation during phototimer radiography. In other words, this abnormality cannot
occur while the phototimer is not provided. If this message is displayed while the phototimer is
not provided, the initial setting may be wrong or the X CONT-2000 PCB may be abnormal.
- 49 -
TSFE-X01010
10.
MUX-10 Block Diagram
INPUT PCH-99
CIRCUIT
PROTECTOR LF
POWER
SOURCE
JINP
X CONT-2000
JINP
NEXSC
JM30
CHON
MGRSON
LV
VP1
PRINTER
J4
STON
MGR-S
CN1-1
VP1
JT1IN
LV
J135
TRANS
T1-MUX10
CN1-2
JPS
13V
JM2
J21
15V、5V
JINV
filament
heating
circuit
VC
to
starter
VC0
STOK
TH1
TH2
V+
control
resistance
SL2
interface
to sheet
panel
J9
AD,BC
IGBT
HAND SW
AR1,HX1
RC RECEIVER
X-RAY,READY,LAMPU
T1,T2
TKV
TMA
STON
JSSR
starter
DOOR CL
REMOTE
CN7
JREM
charge control
JFB
SSRON
CN8
JSP
JM3
CHON
JR12
SL3
SHEET PANEL MUX-A
JSP
V-
MGRSON
CN3
IGBT
DRIVER
CL,C0
CN2
SL1
AR1,HX1
FET
DRIVE
Q14,Q23
RSM1,RSM2
charege
detection
RELAY
CIRCUIT
TKV,TMA,IF
JINV
JSP
125V
JMU
KVBR5
TKV
OVER
-15VOK
JM2
JSP
STOK
LV
INVERTER UNIT-MUX
135V
280V
125V
100V
CHON
J22
MG-D
J280
CN1
DC power
supply
PS1
CN2-2
THERMAL(TH1)
J24
24V
RECTIFIER
13V
135V
ARM SW
DC power
supply
PS2
CN2-1
CN3
CN15
ARM SW
JAEC
JSOUT
CN7
KVFB+
KVFBN1
N2
X,Y,Z,TH1,TH2
SL4
COLLIMATOR
CN15
X-ray tube
H.V cable
13K,0K
LAMPSW
0V
- 50 -
H.V trans
DOOR1
DOOR2
REMOTE
CONTROL
CN4
SW4
ARM POSITION
MUX AEC I/F
PHM
PH-REF
EXR
FSD2
IPOF
PADDLE
SIGNAL
PADDLE
TSFE-X01010
Description of major signals shown in the block diagram in the previous page
Signal Name
Description
AD/BC
IGBT driving signal
AR1
Ready signal
CHON
Charging start signal
CL,C0
Filament heating signal
DOOR CL
Arm accommodation confirmation signal
EXR
Exposure signal
FSD2
Number of lighting fields selection signal
HX1
Exposure
IPOF
Phototimer exposure shutdown signal
KVBR5
Discharge detection signal
KVFB+
+ side of measured tube voltage signal
KVFB- side of measured tube voltage signal
LAMPU
Collimator lighting signal from remote control unit
LV
Transformer T1 primary voltage confirmation signal
MGRSON
Charging suppression resistor reset signal
N1,N2
Measured tube current signal
PADDLE
Voltage signal from phototimer detector
SIGNAL
PHM
Phototimer selection signal
PH-REF
Phototimer reference signal
Q14,Q23
Filament heating FET driving signal
READY
Ready signal from remote control unit
RSM1,RSM2 Filament current signal
STOK
Starter confirmation signal
STON
Starter start signal
TH1,TH2
X-ray tube thermal signal
TKV
Measured tube voltage signal
TMA
Measured tube current signal
T1,T2
IGBT output signal
VP1
Charging voltage detection voltage signal
X-RAY
Exposure switch signal from remote control unit
-15VOK
-15 VDC confirmation signal
- 51 -
TSFE-X01010
11.
Explanation on Each PCB
11.1. NEXSC-MUX10
Function
This is the CPU PCB which controls the MUX-10 using the software. Different from the
NEXSC PCB used in other generators and fluoroscopy units, this NEXSC PCB is not equipped
with an optical cable connector. (The NEXSC PCB equipped with an optical cable connector can
be used in the MUX-10.)
The following data is stored in this PCB. It is recommended to back up the following data in
preparation for PCB replacement.
Initial setting data
FVR value
Phototimer tube voltage correction value
APR
Number of times of exposure
During PCB replacement
When replacing the PCB, set the DIP switches first, then initialize the PCB. After that, perform
the initial setting, adjust the FVR, adjust the phototimer, and register the APR.
Lithium battery replacement
When the life of the lithium battery is expired, the information (contents displayed just before the
power was turned off last time, the error log and the time setting) stored in the SRAM is erased.
When the message requesting battery replacement is displayed, replace the battery using the
following procedure.
① Turn on the power. ON → ② Replace the existing lithium battery with a new one. → ③
Confirm the operation.
- 52 -
TSFE-X01010
LED list
LED No.
LD1
LD2
LD3〜6
Description
Indicate the status in the 7-segment display area as shown
below.
Flickers while the X-ray tube is released.
Meaning of indication by the LD1 and the LD2 in the 7-segment display area
SS: An input is given from a sheet key.
On: Exposure is being performed.
OF: Exposure is finished.
Check pin list
CP No.
CP1
CP2
CP3
CP4
CP5
DIP switch list
SW No.
2-1
2-2
2-3
2-4
2-5
2-6
2-7
2-8
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
CP Name
System clock
System clock
Vbatt
+5V
DGND
Description
CLK6M
CLK12M
Lithium battery voltage
+5V
GND
Status
Always ON
Always OFF
Always OFF
Always OFF
ON: Measured value display mode
Always OFF
Always OFF
ON: Initial setting mode
ON: EEPROM initialization mode.
ON: Phototimer option is provided.
ON: 200 mAs option is provided.
ON: Remote control option is provided.
ON: Printer option is provided.
ON: Bucky unit is connected.
Always OFF
ON: Adjustment mode
- 53 -
OFF: Normal mode
OFF: Normal mode
OFF: Normal mode
OFF: Not provided.
OFF: Not provided.
OFF: Not provided.
OFF: Printer option is not provided.
OFF: Bucky unit is not connected.
OFF: Normal mode
TSFE-X01010
11.2. X CONT-2000 PCB
CP24〜CP31
VR1,VR2
VR3
CP3〜CP17
CP1,CP2
CP32〜CP40
LD2〜LD4
LD1
CP18〜CP23
Function
This PCB equipped with the following circuits performs general control.
1. I/O circuit, A/D circuit and D/A circuit to transfer data between the NEXSC PCB
2. Circuits to detect and feed back the TKV and the TMA
3. TKV OVER detection circuit (discharge detection circuit)
4. System timer circuit (X-ray cutoff signal)
5. IGBT drive control circuit
6. Filament heating FET drive control circuit
7. Dynamic display control circuit, sheet key input circuit and sheet key search circuit
8. -15V detection circuit
9. Power ON reset circuit (master reset signal)
10. Charging control
11. Interface circuit for the Bucky unit
During PCB replacement
The VR1 (mAs), the VR2 (VFADJ) and the VR3 (PULSEADJ) should be adjusted.
Refer to “Appendix C2 X CONT-2000 Adjustment Method” in the Installation Manual.
- 54 -
TSFE-X01010
LED list
LED No.
Signal Name
LD1
KEY IN
LD2
AR
LD3
HX
LD4
CHON-O
Description
Lights when an input is given from the
panel.
Lights when the READY switch or the first
step of the hand switch is pressed.
Lights when the X-RAY switch or the
second step of the hand switch is pressed.
Lights during charging.
VR list
VR No.
VR1
VR2
VR3
Check pin list
CP No.
CP Name
CP1
+5V
CP2
GND
CP3
CHECK1
CP4
CHECK2
CP5
VSIN
CP6
CP7
IFST
IFV
CP8
VS
CP9
TKV
CP10
VN
CP11
CP12
CP13
CP14
CP15
CP16
CP17
CP18
CP19
CP20
CP21
CP22
TMA
PKV
RMA
FVR
PH-REF
GND
+15V
+5V
GND
VP1
VP2
IC
CP23
LV
CP24
CP25
+15V
GND
VR Name
mAs
VFADJ
PULSE ADJ
Description
Adjusts the mAs clock.
Adjusts the inverter working frequency.
Adjusts the IGBT gate pulse.
Description
+5V
GND
To be short-circuited during adjustment. Shirt-circuit them first, then turn
on the power. Short-circuiting them in the power ON status is invalid.
Refer to Appendix C in the Installation Manual.
15V input during adjustment.
Refer to Appendix C in the Installation Manual.
Filament heating voltage signal (after correction)
Measured filament current (1V/1A)
Inverter working frequency voltage until the tube voltage reaches the preset
value
Measured tube voltage (1V/20 kV)
Reference inverter working frequency voltage after the tube voltage reaches
the preset value
Measured tube current (1V/50 mA)
Preset tube voltage (1V/20 kV)
Preset tube current (1V/50 mA)
Preset filament current (1V/1A)
Preset phototimer reference voltage
GND
+15V
+5V
GND
Charging voltage detection (1V/100V)
Not used
Not used
Power supply confirmation (2V/100V) (which detects 100V output from the
T1)
+15V
GND
- 55 -
TSFE-X01010
CP26
CP27
-15V
VIN
CP28
MASCLK
CP29
F
CP30
mA-IN
CP31
CHGREF
CP32
CP33
CP34
CP35
CP36
CP37
CP38
CP39
CP40
GND
A/D
B/C
KVFBKVFB+
KVT
Q1-4
EXRR
Q2-3
-15V
Inverter frequency voltage
Signal created when the TMA is converted into the frequency by the
V/F converter. This signal (5 kHz/50 mA) is used in the mAs timer.
Oscillation frequency to drive the IGBT. This signal is adjusted by
the VR2 (VFADJ).
Refer to Appendix C in the Installation Manual.
To be short-circuited during adjustment
Charging control reference voltage (equivalent to 330V) which is
approximately 3.3V (330V)
GND
IGBT drive control signal
IGBT drive control signal
Tube voltage feedback signal (-)
Tube voltage feedback signal (+)
Signal which rises when the tube voltage reaches the preset value
Heating FET drive control signal
Exposure signal
Heating FET drive control signal
- 56 -
TSFE-X01010
11.3. INVERTER UNIT MUX
Function: This PCB equipped with the following circuits controls the power system.
1. IGBT circuit
2. Filament heating circuit
3. Starter circuit
During PCB replacement
Adjust the FVR again.
Set the DIP switches also.
Check pin list
CP No.
CP Name
Description
CP1
CL
Filament heating voltage
CP2
CO
CP3
Gc
IGBT gate signal
CP4
Ec
CP5
Ga
IGBT gate signal
CP6
Ea
CP7
YC
Stator coil current detection (1V or more is normal.)
CP8
Gd
IGBT gate signal
CP9
Ed
CP10
Gb
IGBT gate signal
CP11
ZC
Stator coil current detection (1V or more is normal.)
CP12
Eb
IGBT gate signal
CP13
GND
GND
Fuse
Fuse Name
Description
F1
Provided for protection
DIPSW
SW No.
Function
SW1A
ON in the MUX-10
SW2A
Not used
- 57 -
TSFE-X01010
11.4. INPUT PCH-99
Function
This PCB equipped with the following circuits controls the power supply.
1. Collimator lamp voltage output circuit
2. Electromagnetic lock voltage output circuit
3. Charging voltage detection circuit
4. Supply voltage confirmation
During PCB replacement
The VR1 (VP1ADJ) should be adjusted.
Refer to “Appendix C3 Charging Voltage Adjustment Method” in the Installation Manual.
LED
LED Name
LED1
Description
Lit while electric charge remains. Check this LED during a work.
VR
VR No.
VR1
VR Name
VP1 ADJ
Description
Adjusts the charging voltage detection function.
- 58 -
TSFE-X01010
Check pin list
CP No.
CP1
CP2
CP3
CP4
CP5
CP6
Fuse list
Fuse No.
F1、F2、F3
F4
F5
F6
F8
Signal Name
VC0
VC
+15V
GND
VPA
VPB
Description
Charging voltage
+15V
GND
Provided for charging voltage check.
VP1=0.246×(VPA−VPB)
Description
Not used
280 VAC line from the transformer T1
13 VAC line from the transformer T1
125 VAC line from the transformer T1
100 VAC line from the transformer T1
11.5. SHEET PANEL
This PCB is used also in the MUX-100.
During PCB replacement
Adjust the buzzer sound volume trimmer.
- 59 -
TSFE-X01010
11.6. MUX AEC I/F
VR1
VR2
Function
This PCB is provided for the phototimer option. This PCB compares the paddle signal from the
detection unit with the reference voltage PH-REF from the NEXSC PCB, and outputs the
exposure cutoff signal IPOF to the X CONT PCB.
During PCB replacement
Adjust the VR1 and the VR2 in accordance with the description on installation of the phototimer
option in the Installation Manual.
VR list
VR No.
VR Name
Description
VR1
OFFSET
Adjusts the offset.
VR2
GAIN
Adjusts the gain.
Check pin list
CP No.
Signal Name
Description
CP1
IN1
Phototimer signal 1
CP2
IN2
Phototimer signal 2
CP3
REF
Phototimer reference voltage
CP4
IPOF
Phototimer cutoff signal
CP5
GND
GND
- 60 -
TSFE-X01010
11.7. RC RECEIVER
Function
This PCB functions as the receiving side of the remote control unit option.
During PCB replacement
There are no adjustment items.
- 61 -
Download