28/06/2022 USER TRAINING BEMS/UT/001 TYPE CODE : 14-361 Ventilators, Intensive Care, Neonatal/Pediatric (Draeger Babylog 2000) Prepared by : Nurhafizah BEMS CONTENT 1. EQUIPMENT OVERVIEW 2. BASIC FUNCTION (not to include clinical application) 3. DAILY AND ROUTINE MAINTENANCE 4. SAFETY PRECAUTIONS IN OPERATING EQUIPMENT 5. CAUSES AND CORRECTIVE ACTIONS 6. CLEANLINESS AND DECONTAMINATION AWARENESS 7. MAINTENANCE REQUEST PROCEDURE 8. POTENTIAL HAZARD 9. AWARENESS ON VIGILANCE REPORTING 1 28/06/2022 1. EQUIPMENT OVERVIEW 11 Front 1 Rotary switch for setting the CPAP/IMV/IPPV operating mode 0 =OFF 2 Buttons for manual inspiration 3 Seat for O2 sensor housing, sealed with cap 4 Connection sleeve for pressure measuring line 5 Inspiration port 6 Expiratory valve 7 R button to reset the alarm after eliminating the cause of the alarm 8 Button G toggles the alarm tone for 2 minutes 9 LEDs 10 Rotary knob to adjust the inspiratory O2 concentration from 21% to 100% 11 Display for ventilation frequency f and lower pressure Pmean • For CPAP, only Pmean is displayed 12 Rotary knob for setting the upper alarm limit 13 Rotary knob for setting the lower alarm limit 14Yellow LED, lights up with IPPV/IMV 15Yellow LED, lights up during CPAP 16 Analog mechanical pressure gauge for continuous display of inspiratory and expiratory airway pressure 17 Rotary knob for setting PEEP/CPAP 18 Knob for time settings TEX/TIMV; blue scale valid for IMV, gray scale valid for IPPV 19 Pinsp setting knob 20 Knob for the time settings TIN 1. EQUIPMENT OVERVIEW Back 1 connection for O2 (NIST) 2 Connection for compressed air (NIST) 3 Adapter compressed air from NIST to DIN connection 4 Adapter O2 from NIST to DIN connection 5 Electrical supply box 6 12VDC to 24VDC input jack 7 Power cord connector 1 10/120 VAC or 220/240 VAC, optional 2 28/06/2022 2. BASIC FUNCTION 1. Insert expiratory valve Use sterile expiratory valve. Loosen the knurled screws and unscrew the valve block from the device. Tilt the valve block 90° forward Insert the diaphragm into the valve block. The lettering on the membrane must be legible. Place the lower left hole on the valve block on the guide pin for the housing connection. The lettering "Expiration" must be legible. Press the valve block with the membrane onto the housing connection. Tighten both knurled screws 2. BASIC FUNCTION 2. Gas supply installation Screw the air and O2 connection hose to the back of the Babylog 2000*. When supplied from a central gas supply Plug the plug into the wall tapping points. When supplied from bottles Connect the air and O2 connection hose to the pressure reducer of the corresponding bottle. *The compressed gases must be dust-free, oilfree and dry! 3 28/06/2022 2. BASIC FUNCTION 3. Power Supply Internal supply - with integrated 6 V battery External supply Before using for the first time or after a long period of non-use: 1 With DC voltage from the vehicle electrical system: – via 12/24V input socket (max. 30 V input voltage). q Plug the 12/24 V DC connection cable into the input socket. Charge the battery for approx. 10 hours at ambient temperatures above 01C. Plug in the mains plug or 12/24 V DC connection cable. With a fully charged battery, the Babylog 2000 can be operated for around 10 hours. The operating time is reduced by reducing the capacity of the battery as it ages and when operated at ambient temperatures below 0°C. 2 With integrated power supply (for devices with electrical supply box2): The mains voltage must match the voltage range specified on the rating plate, 100 to 240 V AC. Nameplate on the underside of the electrical supply box. • Plug in the mains plug. When both supplies are connected to the electrical supply box 2, the electrical supply comes from the mains before the supply from the 12/24 V DC connection cable. The connection of both supplies is possible and does not cause any interference. 2. BASIC FUNCTION 4. Preparation Use sterile hose system! Insert the catheter connector elbow (refer picture) 1 into the "Expiration" outlet and 2 into the "Inspiration" outlet *Attach ventilation hoses, note hose lengths. Connect Y-piece - Adapter K, 90° 3 Slide the sterile Y-piece into the ventilation hoses. 4 Connect the bellows as a test lung to the Y-piece with the catheter connector. 4 28/06/2022 2. BASIC FUNCTION Connect the pressure measuring line Use sterile pressure measuring line. 5 Plug the hose (0.3 m) onto the spout P. 6 Attach the sterile bacteria filter and position it so that no condensate can penetrate. Plug the pressure measuring line (1.6 m) onto the bacterial filter and 7 Put the rubber angle on the pressure tap on the Y-piece. If necessary, shorten the hose accordingly. Connect humidifier Connect the humidifier according to the associated instructions for use. 2. BASIC FUNCTION 5. OPERATION 1 Set required oxygen concentration. Controlled ventilation IPPV 2 Set rotary switch to IPPV operating mode. 3Yellow LED Q lights up. 4 Set the rotary knobs to "green dots", the device now works with the parameters: TIN = 0.8 s TEX = 1.2 s f = 30 bpm or I : E = 1 : 1.5 • Set ventilation parameters patientPEEP = 2 mbar specifically. Pinsp = 20 mbar • Connect patients. • Set alarm limits for pressure monitoring 5 28/06/2022 2. BASIC FUNCTION Ventilation with plateau (Pressure limitation) 5 Knob »P insp« : Adjust according to the display on the pressure gauge, starting with small values, until the desired inspiratory pressure plateau is reached. 6 Knob »PEEP/CP AP« : set. In the ventilation pattern with a plateau, the Pinsp setting prevents damaging pressures, e.g. if the compliance decreases. Ventilation without a plateau (Time limit) 1 Knob »P insp« : Set approx. 5 mbar above the inspiratory pressure peak. 2 Knob »PEEP/CP AP« : set. 2. BASIC FUNCTION PEEP Setting 3 Knob »PEEP/CP AP« Set to the desired value according to the pressure gauge reading. Ventilation with plateau: The effective stroke volume is reduced. 4 If necessary, increase P max. Ventilation without plateau: The airway pressure is increased by the amount of the CPAP value. Increase Pinsp by this amount. Adjust upper airway pressure limit. Conversion from TEX/ TIN to f or I : E f and I : E can be roughly determined from TIN and TEX using the frequency diagram: 6 28/06/2022 2. BASIC FUNCTION Intermittent Mandatory Ventilation (IMV) For weaning from ventilation with IPPV. Mixed form of mechanical ventilation and spontaneous breathing. The mechanical ventilation strokes ensure the minimum ventilation, in between the patient can breathe spontaneously. The ventilation pattern of the mechanical, mandatory breaths works as in IPPV: with and without a plateau. A positive airway pressure can be applied for spontaneous breathing. The device applies a continuous flow of 8.5 L/min. 1 Turn rotary switch to IMV. 2 At the knob »TEX/TIMV« : the blue scale TIMV is now valid. Gradually increase expiration time. The inspiration pattern is preserved. 3 Use the rotary knob »PEEP/CPAP« to set the positive airway pressure for the spontaneous breathing phases. 2. BASIC FUNCTION Spontaneous breathing under positive airway pressure (CPAP) The device applies a continuous flow of 8.5 L/min. An adjustable end-expiratory pressure is generated with the expiratory valve. 1 Turn rotary switch to CPAP. 2Yellow LED w lights up. 3 Use the »PEEP/CPAP« rotary knob to set the positive airway pressure. Set lower alarm limit 4 The displayed pressure Pmean should match the set PEEP/CPAP. 7 28/06/2022 2. BASIC FUNCTION 6. Alarm limit Set alarm limits for IP V/ IMV Lower alarm limit: 1 Set the rotary knob to approx. 5 mbar above the end-expiratory pressure. 2 Green LED P flashes at the beginning of each inspiration = correct setting of the alarm limit. Upper alarm limit: 3 Set the rotary knob approx. 10 mbar above the maximum pressure Pmax. 2. BASIC FUNCTION Set alarm limits for CPAP With CPAP =0, monitoring of the lower alarm limit is not possible. The device warns constantly. Use the patient monitor to detect an apnea! Lower alarm limit: 4 Set the rotary knob 2 to 3 mbar below the endexpiratory pressure. Upper alarm limit: 5 Set the rotary knob approx. 10 mbar above the maximum pressure Pmax. 8 28/06/2022 2. BASIC FUNCTION Start inspiration manually Possible in all operating modes 1 Set the »Pinsp« knob to the "green dot" approx. 20 mbar. 2 Set rotary knob »TIN« to "green dot" or to patientspecific value. 3 Briefly press the »Man Insp« button. After the set inspiration time TIN (with IPPV/IMV) has elapsed, expiration or CPAP takes place. For a prolonged inspiration: 3 Keep the »Man Insp« button pressed for the time of inspiration. 2. BASIC FUNCTION After connection of the patient: 4 Set rotary switch to 0. Set the switch of the additional devices to 0. Pull out all electrical and pneumatic supply plugs. If the battery is to be charged, leave the electrical supply connected. 9 28/06/2022 3. DAILY & ROUTINE MAINTENANCE After Use: Disconnect breathing circuits Disconnect all hoses from the outlets. Pull the catheter connector elbow out of the outlets. Remove Y-piece. Detach the pressure measuring line from the Ypiece and from the »P« socket. Pull off bacterial filter. Unscrew the O2 sensor housing and remove the O2 sensor. Clean according to the relevant instructions for use. At the expiratory valve: Loosen both knurled screws, remove the valve block and take out the diaphragm. 3. DAILY & ROUTINE MAINTENANCE Routine maintenance Bacterial filter for pressure measuring line Replace after 20 times sterilization Diaphragm of the expiratory valve Replace every six months Device inspection and maintenance Every six months by experts Recurring safety checks Every six months by experts Battery pack Replaced by experts after 2 years Major overhaul of pressure reducers Every 6 years by experts Membrane keyboard If damaged, have it replaced by experts 10 28/06/2022 1. Follow the instructions 4. SAFETY PRECAUTIONS IN OPERATING EQUIPMENT Any handling of the device requires precise knowledge of and compliance with these instructions for use. The device is only intended for the use described. 2. Maintenance The device must be inspected and serviced by specialists every six months (with a report). Repairs to the device only by specialists 3. No operation in potentially explosive areas The device is not approved for operation in potentially explosive areas 4. Monitor ventilation Have qualified medical personnel monitor the patient during ventilation 5. Have manual ventilation equipment ready If there is a noticeable error on the ventilator, e.g. If, for example, the compressed gas supply fails, the life-sustaining function can no longer be guaranteed, the patient must be ventilated immediately with an independent ventilation device 6. Do not use mobile phones within a distance of 10 meters from the device Cellular phones can interfere with the function of electromedical devices 5. CAUSES AND CORRECTIVE ACTION Alarms 1 LED J for power supply Green = operation Red = operating voltage low alarm 2 Red LED MIX lights up in the event of a lack of gas alarm 3 Red LED W for upper airway pressure alarm limit 4 Red LED w for airway pressure lower alarm limit for CPAP 5 LED P for fail to cycle alarm for IPPV/IMV Green = inspiration phase for IMV/IPPV Red = fail to cycle (FTC) alarm, inspiration phases are not recognized 11 28/06/2022 5. CAUSES AND CORRECTIVE ACTION In the event of an alarm Alarm tone sounds, the glowing red LED indicates the cause of the alarm. Suppress alarm sound: 1 Press button G. The alarm sound is switched off for approx. 2 minutes. Yellow LED in button lights up. When the cause of the alarm has been eliminated: 2 Press the R button. "Reset" is possible with FTC alarms, even if the cause of the alarm has not yet been eliminated. After a delay time, the alarm is triggered again. When the alarm sounds: • All LEDs go out, monitoring is no longer possible. The alarm tone cannot be suppressed. Then: • Connect to external power supply via mains cable or 12 V / 24 V DC connection cable. The monitoring is immediately ready for use again. • Charge battery In the event of a lack of gas alarm: Press button R, the alarm tone is switched off. The LED display only goes out when there is sufficient gas supply. In the event of an alarm from the electrical supply: Red LED J lights up. Recharge the battery. The battery is ready for operation for approx. 1 hour. 5. CAUSES AND CORRECTIVE ACTION Error Cause Action taken Red LED J lights up Voltage supply for monitoring functions too low: battery discharged. Charge the battery. Alarm sounds, all LEDs go out. Battery discharged, no residual capacity available Charge the battery, if the alarm sounds anyway, change the battery Red LED lights up, alarm sounds Supply pressure too low. Pressure hoses not connected. Check which gas triggered the alarm, make sure the pressure is greater than 3 bar. When supplied from bottles, a remaining operating time of approx. 40 minutes is guaranteed. To suppress the alarm tone for longer, press the R button Mixer function disrupted Red LED W lights up, alarm sounds Upper alarm limit Paw exceeded: Increase in pressure in the hose system. Patient breathes against the device. Check alarm limit setting. Check hose system. Check ventilation parameter settings and change if necessary. Press the R button Red LED w lights up, alarm sounds Lower alarm limit Paw undershot for 17 s with CPAP. Leak or disconnection in the tubing system Check alarm limit setting. Tubing Check for tight connection. Press the R button 12 28/06/2022 5. CAUSES AND CORRECTIVE ACTION Error No breathing possible Condensate in the expiration hose Expiratory tubing kinked Exhalation valve blocked Red LED P lights up, alarm sounds Fail to cycle Alarm (FTC) 6. CLEANLINESS AND DECONTAMINATION AWARENESS Cause Paw has not crossed the lower alarm limit (FTC alarm limit) for 17 s for IPPV or for 40 s for IMV Paw lower than the lower alarm limit: Leakage or disconnection in the hose system Paw constantly greater than the lower alarm limit: device error Action taken Empty the expiratory hose Check expiration hose, create free passage Check diaphragm for proper seating Check the alarm limit setting Hose system Check for tight connection. Press the R button Call for service Perkara ini adalah merujuk kepada: “MASTER AGREED PROCEDURE : BEM-TS-004 (INCIDENTS, HAZARDS AND HANDLING HAZARDOUS AND CONTAMINATED BIOMEDICAL EQUIPMENT ) PARA 6.3 (Handling Contaminated Biomedical equipment )” 13 28/06/2022 START 7. MAINTENANCE REPORT PROCEDURE User identify faults during use Perform Corrective Maintenance Decontaminate faulty equipment Perform performance test/electrical safety test Initiates a request to Medivest Complete work order and assessment detail (Emergency call must be made via direct call to the Medivest’s help desk ) Request number provided to the user and generate Corrective Maintenance Work-Order Authorised Personnel to verify and close work order User Verify Update record in ASIS Respond to the request within the specified time frame 9. AWARENESS OF VIGILANCE REPORTING END There is no vigilance report such as Incident Reporting, Hazard Alert & Recall for this equipment from manufacturer and Medical Devices Authority (MDA) as at 2018. 14