Action Entry Dresden-Team Author: Torsten Boehme Created: 01/08/2015 09:14 AM Last edited by: Torsten Boehme Modified: 01/08/2015 11:17 AM Laser: Laser Serial #: Model / Site / Tool: Sitronics / A52 (61577) 61577 Action Category: Troubleshooting Action Date: 01/08/2015 Laser ShotCt.: SR #: ELS-6600 / Sitronics / A52 Scheduled Downtime Action Plan #: Unscheduled Downtime N/A Special Config/Registry Changed Action Description: Status: During processing a batch some wafers have rejected. In batch report window we see a “Dose Error” in front of rejected wafers and “HV/Ep (upper & lower) level of energy too small” in “Warnings” window. open Actions: 1/8/2015 09:16 AM : Torsten - original customer request: - checked history due to problem still reported last year and not closed: xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx HISTORY 10/03/2014 09:30 AM : xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx 04/22/2014 xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx 09/11/2014 xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx 9/17/2014 xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx - main recommendation: 9/11/2014: - not done so far or no information recieved that customer ordered parts xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx Laser report: - data anlysis of costumer Sitronics system A52 Model: ELS 6600 Laser:SerialNo 61577 FEW report FEW report E 73 Gas supply prs high E 73 Gas supply prs high 12-21-14 17:23:11 12-21-14 17:05:35 - E73 error cutomer feedback: - checked last related diags: D 44 F2 supply prs D 45 Ne supply prs D 46 He supply prs 515.4 552.6 448.8 kPa kPa kPa - result: KrNe and KrFNe still above the gas requirements W W W W W W 151 Bad remote HV data 01-05-15 00:15:29 100 Inject orifice cal. 01-04-15 23:51:10 100 Inject orifice cal. 01-04-15 15:22:40 100 Inject orifice cal. 01-04-15 01:40:04 100 Inject orifice cal. 01-03-15 04:24:54 101 CCI Volts too high 12-30-14 09:06:19 W100, possibly related to the FAC supply pressure above SPEC - recommend to adjust FAC supply pressure against MANIFOLD reading to adjust properly Lifetime report Lifetime report automatic detection based on current Configuratin setting: D 140 MN B D 141 LF B D 151 D 153 D 154 D 155 D 156 D 157 D 158 D 159 D 160 D 161 D 162 D 163 D 164 D 165 D 166 D 167 Shot total Front window Rear window Laser chamber Filter Front optics Rear optics WCM module HV pwr supply Comprssn head Commutator Beam splitter Controller F2 trap Shutter Bearing life 13719 M 13963 M real values due to counter adjustment by customer: 13963 M real values due to counter adjustment by customer: 13963 M real values due to counter adjustment by customer: 13963 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 13718 M real values due to counter adjustment by customer: 260.5 cycles 1941 Kcycles 21431 Hr Exp. LT 12000 hours, almost doubled 15249 15249 15249 15249 15004 15004 15004 15004 15004 15004 15004 15004 - all module lifetimes over limit - even C274/275 are violating the safety aspect: C 274 3000 F2TrapMt C 275 3500 F2TrapLf must be set to 2400 must be set to 3000 Maintenance Report Maintenance Report GsXdcrMt GL value = 8766 hours (MENSOR / Transduce calibration chamber pressure sensor) curent reading: C 220 48987 GsXdcrTm - recommend to perform MENSOR calibration for CH/MAN/VAC Energy / Valtage analysis Energy / Valtage analysis GR - HV clipping in single pulses - bad WL sigma G2 - BW decrease and HV increase over gas life G3 - HV mean reached maximum due to the offset needed for scanner HveP calibration D 213 D 214 D 216 D 108 Avg10kVMinC Avg10kEngy Avg10kVlts Energy Sigma 878 9831 899 1.6 V uJ V % WCM report WCM report D 75 Lambda Fine 2326.2 pm D 76 Lambda Coarse 2326.1 pm D 77 Lambda Dscrpncy 0.1 pm D 78 Lambda Error -0.01 pm D 79 Lambda Linewidth 0.37 pm D 80 Lambda Center 310 D 81 Lambda Stat: L D 82 Lambda Video floor 0.30 V D 83 Lambda Video peak 2.98 V D 84 Lambda Fe PCell 1.25 V D 85 Lambda Left peak 1.74 V D 86 Lambda Right peak 1.82 V D 87 Lambda Temperature 27.61 DegC D 88 Lambda dT/dt 0.01 DegC/min D 89 D 90 D 91 D 92 D 93 Lambda Temp comp. -0.17 pm AWR Lambda 248.3271 nm AWR Lambda Error -0.01 pm AWR Linwidth 0.83 pm AWR PCell Dip 43 % - values OK FAC report FAC report D 44 F2 supply prs D 45 Ne supply prs D 46 He supply prs 515.4 552.6 448.8 kPa kPa kPa - E73 error cutomer feedback: - checked last related diags: - result: KrNe and KrFNe still above the gas requirements W W W W W W 151 Bad remote HV data 01-05-15 00:15:29 100 Inject orifice cal. 01-04-15 23:51:10 100 Inject orifice cal. 01-04-15 15:22:40 100 Inject orifice cal. 01-04-15 01:40:04 100 Inject orifice cal. 01-03-15 04:24:54 101 CCI Volts too high 12-30-14 09:06:19 - F2 supply pressure below specifications: needs to be adjusted within specification - N2 supply pressure below specifications: needs to be adjusted within specification Configuration GLIST check Configuration GLIST check (base) D 7 V3.3 BL9 - Gas configs set per cymer request - F2 trap limits are violation current safety instruction stability configuration gas configuration C 130 260 F_FilDPr C 131 2500 LasngPrs C 132 2900 T_FilPrs C 135 45 InjktDPr C 142 54 BledDPrs C 143 44 BledOPrs C 162 120 InjHrMin C 163 300 InjDltaV C 165 50 F2PerHr_ C 166 534 F2PerMSh energy stability C 304 100 SgmaLock C 305 150 SgmaUnlk C 306 300 SgmaPeak C 307 40 ISHldOff C 308 50 ISSgmaLk C 309 250 ISSgmaPk C 310 60 ISSgmaUL C 311 50 ISWvLckB C 331 3 WCntlBnd C 332 8 LineWMax C 333 5984 WLAvWn_L C 334 5984 WLAvWn_U C 335 0 WLAvBQ_U C 336 225 WLVrBQ_U C 337 0 BndW_Lim C 338 0 AvgVarWn C 339 30 WLockWin xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx Action Plan: short term: 1. Gas Adaption: C130 to 280 (is 260) C166 to 650 (is 534) refill fire laser in WSR mode (internal energy, 10mJ, selflock, SL BI = 0.2s, SL BC = 125, Rep rate 2000 Hz) fill out table below right after refill and every day, take GR provide feedback/ datas Average Pulse Energy Output at Internal HV Mode ( ENERGY MODE) Manual Mode ( TRIGGER MODE ) Trigger Mode: Selflock Selflock Burst Count: 125 Selflock Rep Rate: 2000 Hz BURST COUNT:0 HV (Volt) 950 850 750 Rep. Rate (Hz) Continuous 2000 2000 2000 Date/Notes 675 2000 D107 Pulse Energy (mJ) 1000 High Voltage & Bandwidth 100 HV Reading 2. Blower system adaption after gas daption adjust blower motor speed to minimum WL sigma error fire in WSR mode take GR make adjustments to C370 (stepwise minus 50rpm) take GR compare WL sigma charts in FST and leave C370 at minimum WL sigma value long term allready provide september last year: Energy Mode: Internal Sigma (Energy Mode) Selflock Pulse Interval: 0.1 sec ( REP RATE) Energy =10.0 mJ D 108 D 79 =< 2.0% =< 0.5 pm D 83 =< 5.0 V D 84 D 85 D 86 D 40 45 – 65C D 25 A D16 D 155 Chamber Shots(M) D 151 Total Shots(M) Parts list to order: - as already reported in March and April 2014 all consumables are over EOL (end of life) - as next step Mikron should order all EOL consumables and ask for a service to replace the parts LNM order PN S143731 Chamber order PN S107594 OC order PN S05-06168-00 WCM order PN S136616 xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx