SERVICE MANUAL Electrocardiograph FX-2111 4R2043 FUKUDA DEIMSHI CO., LTD. Copyright © 1997 by Fukuda Denshi Co., Ltd. No part of this document may be copied or transmitted in any form without the prior written permission of Fukuda Denshi Co., Ltd. Manufactured in Japan Foreword A We list up here the warning marks used in Fukuda operation and service manuals. When you sen/ice the FX-2111, read this service manual thoroughly and pay attention especially to instructions bearing the following marks: Warning Marks Warning marks used in operation and service manuals and labelled on the instrument have the following meanings: Read them carefully to understand the meanings and make sure of the significance of each particular. A This mark is used to indicate the direct hazards Danger which may lead to the death or serious injury of the person, may wholly damage the instrument, or may cause fire hazard, unless the instructions written there are followed. A This mark is used to indicate the indirect potential Warning hazards which may lead to the death or serious injury of the person, may wholly damage the instrument or may cause fire hazard, unless the instructions written there are followed. A This mark is used to indicate the possible hazards Caution which may lead to a mild or medium injury of the person, may partially damage the instrument or may erase data from the computer. NOTE "NOTE" is not warning instructions but offers information to prevent the person from doing erroneous sen/icing. Other Marks 0 A Notice to indicate general unspecific prohibited matters. Notice to indicate general unspecific caution, warning or hazard. This service manual describes technical information on FX-2111 to aid the service engineer in troubleshooting. The manual is intended to be used by service engineers of Fukuda representatives and authorized technical staff concerned with medical electronic equipment. Description includes repairing and assembling methods of each component unit of FX-2111. For parts lists and diagrams, refer to the Part II of the service manual. The service manual consists of the following nine chapters: 1 . General Description The outline of FX-2111, specifications, controls and indicators are described. 2 Circuit Description Circuit configuration and functions are explained. 3 Troubleshooting . . Troubles vs. causes and countermeasures are described. 4 Maintenance . Procedures to replace the power fuse, ROM and battery and to perform self-test are described. 5 . 6 . 7 . Periodical Inspection Inspection procedures to prevent troubles and ensure safe and complete operation of the instrument are described. Circuit Diagrams Assembly Diagrams 8 Electrical Parts Lists . 9 . Structural Diagrams A Caution 0 . Never remodel Fukuda medical electronic equipment . The service manual . is intended for the service engineers of Fukuda representatives and the technical staff concerned with medical electronic equipment. Servicing, reassembling, and adjustment shall be performed by authorized service engineers. . Prepare proper facilities and tools when servicing . Be sure to follow the instructions of operation manual when operating the instrument. For operating precautions, refer to the operation manual. . ii A Servicing Precautions A Cautions listed below are the instructions of prohibit danger, warn, ing, and caution described in this service manual. When taking the procedure bearing the following mark, read the description thoroughly, then start the task. A Caution (page 3-2) When checking the power supply and related circuits for troubleshooting, take sufficient care to avoid a short circuit. A Caution (page 3-3) When checking power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand. A Caution (page 3-8) The inserting part of key connector is made of carbon. Avoid frequent repeated disconnection and connection. A Caution (page 4-1) When replacing power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand. A Caution (page 4-1) When replacing the ROM, be sure to turn the power off. Also, take care to install the ROM in correct position. A Caution (page 4-2) When replacing the battery, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand. iii A Caution (page 4-6) Disassembling/Reassembling Precautions Be sure to disconnect the power cord and make sure the instrument is turned off before disassembling or reassembling. . Do not remove the battery before disconnect any PC board . Take care that repeated disconnection of the key panel and sensor board may result in poor contact. Use proper tools to loosen screws. . When reassembling make sure that all screws are securely tightened and all connectors are completely inserted. , PC Board Handling Precautions . PC boards are equipped with highly sensitive components to static electricity. . PC boards are highly sensitive electronic devices Put removed PC boards in a proper protective bag or take appropriate measures to protect them. . Handle PC boards carefully. A shock to them may damage the components. Never insert a connector to the powered PC board nor remove the powered PC board. A Caution (page 5-2) If you find a value which exceeds the allowable level, be sure to let the user avoid using the FX-2111. If the user operates the FX-2111 as it is, he/she may receive an hazardous accident. Equipment Classification The FCP-2155 is classified into the following equipment: 1 Protection against electrical shock . Class I 2 . Type against electrical shock Applied part: Type CF 3 Degree of protection against harmful water invasion Other equipment 4 Degree of safety in using under air-inflammable anesthetic gases or oxygen/nitrous oxide-anesthetic gases Equipment used under an environment containing no inflammable anesthetic gases or no inflammable cleaning agent. . . 5 . Running mode Continuous running mode iv CHAPTER 1 General Information 1 Outline of the FX-2111 . 2 . 1-1 Specifications 2 1 Electrocardiograph Section 1-1 1-1 2 2 General 1-2 23 1-2 . . . Environmental Conditions 3 Controls and Indicators . 1-3 1 Top Panel 3 2 Side Panels, Left and Right 1-4 1-4 33 1-4 3 . . . Bottom Panel 1 Outline of the FX-2111 . The FX-2111 is the easy-to-use single-channel electrocardiograph featuring a simple operation panel. The compact yet high-performance design makes it suitable not only for use in the consultation room but also for-carrying in a visit to the hospital ward or patient s home and examination at an emergency site. ' It has a high-density thermal dot printer incorporated to provide clear ECG recording. The FX-2111 can operate on either AC line or rechargeable Ni-MH battery. 2 2 . . Specifications 1 Electrocardiograph Section Input circuit: Floating from the ground Leads: Standard and Cabrera 12 leads Input impedance: Input circuit current: Calibration voltage: 20MQ min. (referred to 10Hz) 5x10"8 max. 1mV within ±5% Common mode rejection: 10mm or less per IEC 62DC06 test method Polarization voltage: ±300mV min. Time constant: 3 2 sec min. Sensitivities: 1/2, 1 and 2 cm/mV . Frequency response: 0 05 AC filter: 50/60Hz, -20dB or lower Muscle filter: 35 to 45Hz, -3dB (-6dB/oct) 10mm/0.5V, unbalanced, 100kfi min. Liquid-crystal display, 20 chars, x 2 lines (character: 5x7 dots) Thermal dot printer, 8 dots/mm DC input: Display: Recording system: Paper speeds: Chart papersRoll paper: Z-fold paper: . to 150Hz (within -3dB) 25 and 50 mm/sec within ±3% 63mm or 50mm wide x 30m long 63mm or 50mm wide x 20m long, 75mm/fold AID conversion: 12 bits Sampling rate: 1ms 1-1 2 2 General . Safety: I EC 601-1 Class I .Type CF . Internally powered equipment Type CF (IEC 601-1) Power requirementsAC operation: DC operation: Dimensions: Weight: 115V AC, 50/60H2; 19VA max. 230V AC, 50/60Hz; 19VA max. 9 6V, 7W max. (rechargeable Ni-MH battery) Continuous operation: approx. 120 min at 20oC (according to IEC 62D Testing Method) Charging time: within 3 hours 26 (W) x 18.2(D) x 6.3(H) cm Approx. 1.7kg (excluding battery) . 2 3 Environmental Conditions . OperatingTemperature: 10 to 40oC Humidity: 30 to 80%RH (no dew condensation) Atmospheric pressure: 70 to 106kPa Transportation & StorageTemperature: -10 to 40oC Humidity: 10 to 95%RH (no dew condensation) Atmospheric pressure: 70 to 106kPa 1-2 3 Controls and Indicators . Top Panel X Liquid-crystal Display 31 - k » @ Paper Magazine (D Operation Panel Left Panel Right Panel o« ® Potential Equalization Terminal 0 0 © Power Connector 0 F3 © Main Power Switch D @ DC Input Terminal Q © Lead Connector Bottom Panel @ o o { * I i< " iii 0 © Fuse Holders §) ~ ~ 0 ® 1-3 0 © Rechargeable feattery Room 3 1 Top Panel . © Display: Indicates current status such as recording mode, lead, sensitivity, heart rate, or program setting. © Operation Panel: ON OFF Mode Turns the FX-2111 on in DC operation. Turns the FX-2111 off in DC operation. Pressing this key during AC operation lets the FX-2111 enter the charge mode. Select a mode in the following order when pressed during cessation of recording: |-> Automatic recording -» Manual recording -> Programming -| Holding this key during automatic recording lets the FX-2111 record the current lead continuously until detached. Holding this key during manual recording lets the FX-2111 record an event mark until detached. Sensitivity Lead Select < Reset 1mV Start/Stop © Paper Magazine: 3 Selects a recording sensitivity. Select a lead for recording. In programming mode, these keys allow you to select a setting value. Resets the measuring circuit while held during recording. In programming mode, this key selects a setting parameter. Applies a 1mV calibration waveform when pressed during recording. In programming mode, this key selects a setting parameter. Starts the FX-2111 recording the ECG waveform. Another press stops it from recording. Accommodates a chart paper. 2 Side Panels, Left and Right . ® DC Input: Inputs external DC signals with a sensitivity of 10mm/0.5V. © Lead Connector Connects to the lead cable. © Power Connector Connects to the power cable. © Main Power Switch: Turns AC power on/off. @ Potential Equalization Terminal: Makes the FX-2111 equipotential to another instrument used in combination with it. To that effect, connect both instruments to a common grounding conductor using an optional grounding wire. 3 3 Bottom Panel . © Battery Room: Houses the rechargeable Ni-MH battery. © Fuse Holders: Have power fuses inserted. 1-4 CHAPTER 2 Circuit Description 1 . 2 . Introduction 2-1 Isolated Input Circuit 2 1 Buffer Amplifier and RF Driver 2-2 2-2 . 2 . 2 . 2 Lead Network and Lead Selector 2-3 3 Preamplifier and 1mV Generator 2-3 4 R-wave Detector and Overinput Detector 5 Amplifier Control 2 6 Signal Isolator 2-4 2-5 2-5 27 2-6 2 . 2 . . . 3 . Power Isolator Middle Amplifier and AID Converter 1 Middle Amplifier and DC Input 2-6 2-6 3 2 AID Converter 2-7 2-8 2-8 3 . . 4 Motor Control 5 Sensor Circuit . . 1 Detection of Magazine Open Condition 5 2 Detection of Paper End and Paper Marks 5 . . 6 CPU Circuit 2-9 2-9 6 2 CPU, ROM and RAM 2-9 2-9 2-9 63 2-9 . 61 . Reset Circuit . . Gate Array LCD Control Circuit 8 Thermal Print Head Control Circuit 2-11 2-11 Memory Backup Circuit 10. Power Supply and Charging Circuit 2-12 2-13 7 . . 9 . 10.1 Introduction 2-13 10.2 Rectifier/Smoothing Unit and ON/OFF Control 2-13 10.3 + 10V and +5Vd Power Generators 10.4 ±5Va Power Generator 10.5 +24V Power Generator 2-14 2-15 2-15 10.6 Charging Circuit 2-16 1 Introduction . The FX-2111 is composed of the following circuits: (1) Main Board PCB-6409 -Isolated input circuit -Middle amplifier and AID converter circuit -Motor control circuit -Sensor circuit -Reset circuit -CPU circuit (CPU ROM, RAM, gate array) , -LCD control circuit -Thermal print head control circuit -Memory backup circuit -Power supply and charging circuit (2) Sensor Board PCB-6239 (3) AC Inlet Board PCB-6410 ECG signals input through the lead connector are amplified by the isolated input circuit and the middle amplifier circuit, then converted from analog to digital signals by the AID converter circuit. The digitized signals are digitally filtered, if the filter is set to ON, then sent through the gate array to the thermal print head control so that they are recorded as ECG waveform by the thermal print head. During these processes, the heart rate detected by a hardware technique and operation status such as filter ON/OFF setting are displayed on the LCD. The power supply provides circuits with necessary powers. The charging circuit, if activated, charges the Ni-MH battery in AC operation or transfers the power from the Ni-MH battery to the power supply. The state of battery is indicated on the LCD. : PCB-6409 i5V - AC Inlet I0V " re O Isolated Input Circuit i If 3 Power Supply -24V AC Power VDO +2 5V . ZEE ' ' DC Input /Charging;-; - AyiCircuil ;-- Motor Motor Control Thermal Thermal ] Print Head LCD * . 5 M mt :' J Gate Array , CPU Circuit Print it Head [ Control 1 LCD Control PCBSensor Board . Backup Circuit 6239 Overall Block Diagram 2-1 NiMH Battery * DC :pcb-64io Key Panel 2 . Isolated Input Circuit The isolated input circuit is composed of the following: (1) (2) (3) (4) (5) (6) (7) Buffer amplifier and RF driver Lead network and lead selector Preamplifier and 1mV generator R-wave detector and overinput detector Amplifier control Signal isolator Power isolator R - 5 I L F CI 2 D 5 5 5 t a5 c - " I ECG - - 1 5 D - . 5 I 2 - C6 5 5 5 - O 5 Preamplifier " > Signal Isolator I 1mV S RESET Amplifier T RF Drive <- " Control Signal Isolator RF Overinput/ (Photo R-Wave Detector Circuit Coupler) I i i .e- Power Isolator <- Block Diagram of Isolated Input Circuit 2 . 1 Buffer Amplifier and RF Driver oac- Hit icr I 0 9 ISO* . 3 5 Since the buffer amplifier should input signals with high impedance and output them with low impedance, it is configured as an impedance conversion circuit using an operational amplifier. The circuit shown above is individually applied to all leads except for RF (right foot). Also, a limiter using a dual transistor is mounted for protection against overinput. The RF driver feeds back the composite signal of limb leads to the right foot for improved common mode rejection ratio. 2-2 2 2 Lead Network and Lead Selector . OK 211 OK *a? i z 10 Ox > - 3 304 OK i 20 1 0* Lead Select Signals It:? ? ft: io«r Leads A BCD STD 0 0 0 0 10 0 0 4; Ok II 0 10 0 III 110 0 Vr 0 200 0-< J I Z0« OK El 10* Tan m l OX t T 0 10 aVL 10 10 aVF 0 VI 1110 V2 0 a ' UNS2 L] 110 0 0 1 V3 10 0 1 V4 0 10 1 V5 110 1 V6 0 0 11 The lead network is formed with resistors and lead selector, with multi- plexors IC36, IC38, IC41 and IC4. Input signals to multiplexors are selected from RA, LA, LF, and CI to C6 according to four signals of A, B, C, and D then synthesized to produce each ECG lead (see table above). , 2 3 Preamplifier and 1mV Generator . C-1 fcTzOOHK o.aou r-II "V c«i 4 cb> X . in* rc .200 10m i JOK f MOK < 4= " [ F _ .on r -*- i J frt5 i i 1 |Slr4i»i (C J7I a ® ? cm. J o CCC *.«« _ OIL 1SS . mi 117 1 1*3 1 . IK t 3C on* r mioi 5: A icia ® 41J LIt004C$l-i 2 . rc 1 Preamplifier Signals output from the lead selector (multiplexors) are amplified 8 times by a differential amplifier, then 25 times by a noninverting amplifier, thereby letting the isolated input circuit amplify the input signals by 200 times in total. Signals output at the first stage are sent to the overinput 23 . . detector. 2-3 The 3.5Mfi resistor and 1|jF capacitor set the time constant at 3.5 seconds. 2 3 2 1mV Generator . . The 1mV generator divides the high-precision voltage reference output of 1.235V by a high-precision resistor, thereby applying a 8mV voltage to the preamplifier when the 1mV application signal is at a low level. 2 . 4 R-wave Detector and Overinput Detector era Q.Qit Li C«2 [C4 1C J * in * . > 10QN 4*2 U*f 111 I 4M @ lOOH nrvr-!> - 0 23 . JOCK c 111 0 . 1J .. »2 CP ft I 4 a*] UMf I LI l a r- C 42« 0 001 J . C12 J "I 20OK re 2 41 . . R-wave Detector RA and LF signals of lead network are synthesized to II lead. The Rwave detector detects R wave by picking up R-wave component from the II lead signal through a band-pass filter, then sending it to a comparator. 24 . 2 Overinput Detector The signal output at the first stage of preamplifier is sent to a com- . parator and if the signal exceeds ±360mV, the output of the comparator becomes inverted. Signals output from the R-wave detector and overinput detector are adjusted in the pulse width by the multivibrator IC40 (TC4538), then input into the photo coupler. +2 9V . +0 31V . OV ov 1T63 - 0 31V . - 2 9V . TP19 * 170»S TP20 ITOins R-wave Detection Wavelorm Overinput Detection Waveform 2-4 5 Amplifier Control 2 . * - " C C S . IK <5id* 2 £ n 5 of3 Si a IT 1 ® . 2 17 IK HJU 1 . ~ 1 XL lO.Ot - v 0 CO This circuit picks up a reset, 1mV,or lead select signal from the serially transferred amplifier control signals. !<-% 1ms ; I 1 KH 2 AMP C L K AM P - DATA DO D7 D6 D5 D4 D3 D2 Dl DO D7 D6 D5 D4 D3 D2 DL DO D7 DO i i OLD DATA <- > NEW DATA - i i 8ms - 5 Data wfitlen (rom CPU are latched. ? 0 a 0 8ms Response K time: 32ms + a max. _ 16ms + a mm. 2 . 6 Signal Isolator CT* IC21 rjJiSJLp-Iilp t| 1 o.c 0+? . am <fl rCLK H U U « -cTTirClK 1 C 73 lM O frwrr ICJW y5)*»3L. l JLj-l1| IK mil n» Ik --W - l 0 c -; >r i 111 * - t 12» LTT- T1 en CE3d> - r 41 0 01 . i _ aiir _ ___ _ p y j- v -- -tret? > O.Ol o.ot 1 BIO- rLji_. a) i Cr- 1 ISI«3) sr: i 4 The signal isolator modulates input analog ECG signals with FETs Q31 and Q32 and pulses TRF1 + and TRF1 -, then transfers the modulated signals through the T1 isolation transformer (1:1). FETs Q22 and Q23 demodulate the signals. A low-pass filter formed with capacitors and resistors in two stages eliminates noise in the demodulated signals. Digital signals (AMPCLK, AMPDATA and HR) are transferred using a photo coupler. 2-5 2 7 Power Isolator . -020 <ianz> 11 <3B£J-pW-i 0 s ISV , . *fsl\v . iSiiQl f -r- 1 30 ' r JC 10 * i '2 JH 5p TVrr-J ..7,11V T r 11 ISSIOJ <2irD-«-f-1 J- T 1 « ilO -r- *a . , lisjoj T 0 . 1 C 10 200xhi The power is transferred using 10V pulse at 100kHz. At the isolated side, a 3-terminal regulator supplies ±5V. 3 3 . . Middle Amplifier and AID Converter 1 Middle Amplifier and DC Input C4 T i _ £» °- :v.\] at i 3; * imrraT> 1 10 .0.2 -» il-Lt irrw 1 . LA 1 0 21 . > " 10. ? ... ECG signals demodulated at the isolated signal transmission section are output to the middle amplifier just after passing through a low-pass filter formed with capacitors and resistors in two stages. The middle amplifier amplifies the input signals 1.25 times to a total gain of 250 times and sends them to the AID converter. Signals output from the AID converter are digitally filtered with a software technique. The isolated input circuit and the middle amplifier are designed so as to provide a general hardware frequency response of 150Hz (-3dB). The DC input amplifies input signals 0.5 times, then sends the signals to the AID converter through a buffer amplifier. 2-6 3 2 AID Converter . icu. ... it:? DC Li J .» 100 T * 0.022 ' [l 01 l« " * I I -3 '»» t<S302 I Tin 100 T£l« 4 JQ» 1C4l«4 I|l 010* ittlQI 0 O22 . 1 C*1 t 1 HP 100 * 0 1300f ,1 El00" L vvv-r 0110 7 5 w- r -<} a it The A/D converter is a single slope type. It is composed of a triangular wave generator, sample & hold circuit, and comparator. The comparator compares input signals with ±3.2\/ triangle wave generated at every 1ms. The signals are then converted from the voltage to a pulse width and transferred to the gate array. In the gate array, the pulse width is counted, then converted into a digital value. 520ps < 5k Si 480,1s A/D CLK --<? > 96ps ;< 96)13 S/H (TP35) Bps +3 2V . OV Triangular (TP7) 3 2V - . Start ol counting A/D Conversion Timing Besides ECG signals, the AID converter digitizes DC input signals, battery voltage, thermistor signals of thermal print head, and the reference voltage of 2.5V for calibration of A/D conversion. 2-7 4 Motor Control . </00 . LJJ s nn? > @ .« (IT T lOQ- LI J ® 11 - - C* VCOIm - - i Li C3 Li 11% _ J J Li 11 Li * t J* 10* _ 4 itsio: I - . T jiv ; \ Ten 14 1 I C «1 The motor to drive the recorder is the DC motor which has a photo sensor built in for detection of motor speed. The PLL control IC25 compares a signal detected by the photo sensor with the reference frequency in phase, thereby providing a motor control signal at pin 13 of IC25. The motor control signal goes to the integrator circuit formed with R140, R142, and C64 and is made by a proper motor drive voltage of the PWM control IC24. 25mm/s (256Hz) SOmra/s (5L2Hz) r i n ~ r~i m -+5V IC25 (pin 14) SIN - - ov +5V IC25 (pin 3) CIN .OV .+5V IC25 (pin 13) . +3 IV . P2 \ K Delayed Early +2 2V . \ Delayed / Delayed OV Early Delayed 5 Sensor Circuit . The sensor circuit is provided to check for a magazine opening condition and paper end as well as detecting paper marks. 1*4 0» ® fi£2£E2> T t la 7 1 lit : > il 3£ i & <1 if* - llflOl 2-8 4 n* p re it© -H> - t*ii I 51 . Detection of Magazine Open Condition When the paper magazine is open, the microswitch SPVC2-1 is turned off to send an interrupt signal to the gate array. Magazine opening is detected with the fall signal and magazie closed is detected with the rise signal. 52 . Detection of Paper End and Paper Marks A reflection type photo interrupter acknowledges a black paper mark or paper end status (no paper remaining) if receiving no reflection from the location of paper. It discriminates between paper end status and paper mark based on a duration of the low level starting from a fall signal. 6 CPU Circuit . The CPU circuit is composed of a reset circuit, CPU, ROM, RAM, gate array, and backup power circuit. 61 . Reset Circuit VOO 1 14 10K ION vsa vACSEt * ViB 12 !2M * VSC K h J OUIC - QhO Cf 0 C%3 - k cj Q-i L'ltV M O T . . . Using NJM2103, the reset circuit generates a reset pulse as well as monitoring the battery voltage. It outputs the reset pulse when Vdd falls to lower than 4.27V or when it receives a signal from WDT. As for the battery voltage, the reset circuit monitors the power voltage divided by resistor and when the voltage becomes lower than 7.5V referred to the battery terminal, it outputs a pulse to the LOWS terminal to turn the power off. To protect the output pulse against power fluctuation noise, a filter formed with a capacitor and resistor in one stage and a Schmidt trigger inverter are provided. 6 2 CPU, ROM and RAM . The CPU is HD63B09E. It controls the overall circuit of the instrument through the gate array. The ROM is 64K-byte AM27C512-150DC and the RAM is 8K-byte SRM2264LM12. By changing the jumper connection, the ROM can be replaced with a 128K-byte ROM and the RAM, with a 32Kbyte RAM (with a data holding current of lower than 2\iA). 6 3 Gate Array . The gate array is FD88007-AC with a clock frequency of 16MHz. Its control signals are as follows: 2-9 631 . . 2MHz E and Q Clocks for CPU ' -5 00 n s--> < E 500ns V, $->, 125ns Q 63 . 2 Interrupt Signals . There are three types of interrupt signals-NMI, FIRQ and IRQ. All these signals, active at low level, are output to the CPU. -NMI Output at every 1ms. -FIRQ Output at every 10ms. -IRQ Output when any one of MAGAZIN, MARK, HR, LOWB, CHG IRQ and AC/DC signals is input. , !< S> Ims NM I - j* 500jis 1 0 0 Hz 1 Oras FIRQ . CLEAR MARK RUN a PAPER END a = 390ms (9.75mm) with a paper speed of 25mm/sec 190ms (9.75mm) with a paper speed of 50mm/sec NOTE: ACDC, CHG IRQ, and MAGAZIN DOWN signals are detected at their rise and others are at their fall. 63 . . 3 Watchdog Timer (WDT) . 10ms , ' < =» 1 0 0Hz WDT CLEAR , 70ras . WDT a = Response time of IC 2-10 a i 634 . . Chip Select Signals Chip select signals are all active at low level. WE: Write enable signal RD: Read enable signal E Q ROM ] RAM [ KEY J 1 1 | 1 WE I RD J ' 1 ' ' NOTE: LCD signal differs from the above since the memory is ready. (See LCD control circuit.) 7 LCD Control Circuit . Since the LCD (NDM202A00) has a controller built in, it is interfaced directly with the CPU. However, E and Q clocks make the bus timing as shown below. E Q LCD Data Bus Address Bus CM ZX $3040 ~ $30<tF > X 8 Thermal Print Head Control Circuit . A waveform is traced at every 1ms and alphanumerics are at every 6ms with a paper speed of 25mm/sec or every 3ms with a paper speed of 50mm/sec. The RAM for the thermal print head control uses 4K-byte for alphanumeric data. This circuit controls dot heating temperatures based on previously printed data to provide a proper printing condition. When the paper is not driven or when the magazine is open or there remains no paper in the magazine, the circuit generates TOFF signal to shut off power supply to the thermal print head at the FET Q24, thereby protecting the thermal print head. There are four control signals of CLOCK, DATAIN, LATCH, and STROBE. The STROBE signal is varied in pulse width according to thermal print head temperatures. 2-11 1ms LATCH e- I 8,3 STROBE us - !< - 960vs (Max) 895,is CLOCK (SOOKHz) 9 < . 5us S DATA 2v s CLOCK . 1 . Sks 1 DATA 9 . Memory Backup Circuit The battery-backed RAM SRM2264LM12 is installed to keep program contents when the FX-2111 is turned off. The RAM should feature a data holding current of lower than 2nA. J 2 k otci 1 . o> O . I T ." 5 " 9 When the FX-2111 is turned on, Vdd is supplied to the circuit and turns Q6 and Q7 on to send +5\/ Vdd to the RAM. When the FX-2111 is turned off, Q6 and Q7 are turned off but the back- up lithium battery supplies +3V to the RAM to let it keep the stored data. 2-12 10. Power Supply and Charging Circuit 10.1 Introduction The power supply transforms the AC power by the transformer or receives DC power from the battery, then regulates the power to stabilized voltages required by respective circuits. The FX-2111 is equipped with a charging circuit to permit the user to charge the Ni-MH battery. AC Power Transformer Rectifier 5 - + 1 0 VN Regulated Power Supplies + I 0 V ON/OFF Control DC Power + 5 VD Battery I ! ± 5 VA Charging + 2 4 V Circuit Block Diagram 10.2 Rectifier/Smoothing Unit and ON/OFF Control 10.2.1 AC Operation The AC power voltage is transformed by the transformer, then rectified by D1 to D4 and smoothed by C136. D1 to D4 are Schottky diodes with low Vpand form a bridge circuit. This rectifier circuit suppresses voltage decrease and features less heat generation. 10.2.2 ON/OFF Control In battery operation, a press of the ON key switches FET 04 on and electrifies D7, thereby making the base of 05 driven. If the operator detaches the finger from the ON key, 04 is kept on and the power is supplied to the FX-2111. The battery voltage is detected via D6. If the voltage between battery terminals is lower than 7.5V or so, the gate array generates POWER OFF signal to turn 03 on and cancel the bias voltage of 05, thereby switching 04 off. As the result, the FX-2111 receives no power supply and is turned off. Similarly, a press of the OFF key cancels the bias voltage of 05, thereby turning the FX-2111 off. In AC operation, 01 is on; therefore the base of 05 cannot be driven and 04 cannot be kept on, thereby making battery operation impossible. 2-13 VDO AC DETECT D 1 - « rpBHogy -(sczeo r i ' C t 36 if" H ' - >fou€R {>-' 0? tJNS2 1 2 OS D3FS«A in or 3-2P - 20SA gnr Tr fr T-Tr -rrTinr " T - Shis T w It E32EHZ> - - 2(> -O- «25 tOOK HI7 ) 20OK < azcz-S.iv " F 12 UMS212 - K2- <\ LSS301 0- UNS2 1 Z 08 T P48 UNS2 12 {> «2fc 20K EKZO - ear t -> 0 o T I 10.3 * - +10V and +5Vd Power Generators 10.3.1 10V Power Generator The dropper type 4-terminal regulator IC22 is used to regulate unstabilized IOVnU) stabilized lOV. R118 and R119 are to set the output voltage. The 10V power is supplied to the motor and the 5Vd power generator. If the battery voltage becomes lower than 10V in battery operation, IC25 drops the voltage to several ten millivolts. 10.3.2 5Vd Power Generator C7 is a drop type low-loss 3-terminal regulator. It produces regulated 5V power from 10V. IC7 can operate if an input-output voltage difference becomes up to 0.5V and thus it can output the stable 5V power if the input voltage fluctuates due to lowered battery voltage. The 5V power thus produced is supplied to the CPU and other digital circuits. IPS* IC22 IC7 POIQIV11 iovn -> - I TP27 U- msz - 11? 3 C20 O.Ol O . [ CIS I 3 119 « JO gno > 2-14 * TP38 OFT 10.4 ± 5Va Power Generator Using the unregulated 10Vn power from the AC power supply or the battery, the 3-terminal regulator ICQ produces +5V. IC10 is a CMOS inverting type switching regulator, in which the output voltage is fixed and which inverts the input +5V and outputs -5V. A capacitor at the output is an organic semiconductor capacitor which features superior characteristics at low temperatures and least leakage current. ±5V powers thus produced are supplied to analog circuits such as operational amplifier. ICS GEC - .Not hj n« C I J D Ol ZZ. . VIN -VOUI oT* ir 44* ss , - C2» ® 10.5 ft > +24V Power Generator Using the unregulated 10Vn power from the AC power supply or the battery, the +24V power generator, a boosting type DC-DC converter, produces 24V power for driving the thermal print head and charging the battery. IC6 is a PWM-controlled switching regulator. R22 and C18 determine the oscillation frequency. The PWM control is performed at approximately 170kHz to drive Q9 at a high speed. While Q9 is on, the current flows to LI which in turn stores the energy. When Q9 is turned off, the stored energy becomes a counter-electromotive force and is stored in C137 via D13. By repeating a series of these operations, the +24V power generator can produce an output voltage larger than the input voltage. R41 and R42 set the output voltage at 24V. R24 checks for an overcurrent and a characteristic is provided to gradually lower the output voltage when the load current increases to over 0.4A. The 24V power thus boosted is used to drive the thermal print head. When the thermal print head is printing, a signal from pin 21 of the gate array drives the base of Q15 to switch FET Q24 on so that the -I-24V is supplied to the thermal print head. When the thermal print head is not printing, Q15 and Q24 are off and the thermal print head is not supplied with power. Pin 16 of IC6 outputs 2.5V, which is used by the AID converter and the thermistor temperature detector of thermal print head. I - 11 ;; j1UY i 111 * 51 i = LIL _ 2-15 1 UJ 1 ;v J .. i-0-<nm 10.6 Charging Circuit The charging circuit is equipped with the quick charging IC18 (bq2003), which performs constant current control as a frequency modulation controller for switching regulation of the charging current and monitors the battery temperature and voltage and the charging time to ensure proper charging. It also checks whether or not the Ni-MH battery is connected and enables charging only when the battery is connected. In AC operation, the circuit is supplementarily charging the battery even if the FX-2111 is not placed in charging mode. 10.6.1 Starting the Charging Circuit When the OFF key is pressed in AC operation, Q11 is turned on to supply +5V for operation of the IC. Since the base of Q12 is driven by turning Q11 on, Q11 is kept on. When +5V is supplied to IC18, the charging circuit automatically starts operating. 10.6.2 Charging Temperature and Voltage Monitoring/Control Charging is limited by battery temperature and voltage so that it is made in a preset range. To monitor the battery temperature, a thermistor (with negative temperature coefficient) is installed in the battery pack. The thermistor outputs a voltage signal to pin 6 (TS) of IC18 and charging is enabled if the voltage is within the preset limits in voltage converted from battery temperature. The sensitivity of a battery temperature rising ratio (AT/At) is adjusted by resistances of the thermistor and R59, R61, and R62 on the charging circuit. The battery voltage is divided to a voltage per cell by R58 and R60 which are connected between battery terminals, then sent to pin 7 (BAT) of IC18. Charging is enabled if the voltage is within the preset limits in voltage per cell. Divided +5V powers of R84 and R85 are used to determine the maximum cell voltage (MCV) of the battery and the maximum voltage (TCO) for the battery temperature to stop charging, respectively. They are output to pin 10 (TCO) and pin 11 (MCU) of IC18. Preset battery temperature limits and cell voltage limits are as follows: . Battery temperature limits Minimum charge enable temperature limit LTF: Approx. -50C (voltage level 2.0V) Maximum charge enable temperature limit HTF: Approx. 60oC (voltage level 1.13V) High temperature to stop charging TCO: Approx. 650C (voltage level 1.01V) . Battery cell voltage limits to enable charging Minimum: Approx. 1.0V Maximum: Approx. 1.78V That is, when the charging circuit is turned on, it starts charging the battery if the battery temperatre is within a range of 50C to 60oC and the cell voltage is within a range of 1.0V to 1.78V. If either one of the above 2-16 conditions is not satisfied, it does not start charging the battery but is placed in a standby status until both conditions are satisfied. Once the charging circuit starts charging the battery, the charge complete system described later will be effective. However, if the maximum cell voltage of 1.78V is exceeded before the start of charging, the charging circuit judges the battery abnormal and does not start charging. 10.6.3 Charge Current Control The power required for charging is supplied from the 24V power generator. The charging current to the battery is detected by the resistor R18 then sent to pin 9 (SNS) of IC18. The output of pin 14 (MOD) of IC18 is switched based on the voltage value of R18 to switch Q17 on/off, thereby making the current constant. The charging current is set at approximately 0 45C (480mA). . Q16, D22, and Q10 form a MOS FET driver and level shifter circuit to heighten the switching frequency. 10.6.4 Charge Complete System When the battery temperature and cell voltage are within the respective ranges, the charging circuit continues to charge the battery until one of the following five conditions is satisfied: . - AV detected . . . . - AllAt (temperature rising ratio) detected Maximum temperature of 650C to stop charging TCO reached Maximum charging time of 3 hours reached Maximum cell voltage MCV of 1 78V reached . Usually, the charging circuit stops charging by detecting -AV. However, if charging is started at high ambient temperatures, the charging circuit may detect the maximum temperature (TCO) to stop charging. Also, as a safety measure against no - AV detection, the maximum charging time of three hours is set by timer and TM1 and TM2 signals at pins 4 and 5 of IC18, respectively, stop the charging circuit from charging if the time is reached. ...... jLL1Ji - «_ s I ' - 7 A (tl " - Hit © 1 , r s net . <1 U l. ix r - 5 1 I T.. 2-17 T -S r 10.6.5 Other Functions In AC operation, supplemental charging is made by R66. In AC operation, Q14 is on and the base of Q13 is driven to supply R66 with the 24V power. The current of which the value is determined based on a voltage difference at R66 flows for supplemental charging. However, if the battery is not connected, Q8 is turned on and Q14 is kept on, thereby disabling supplemental charging and at the same time cancelling the bias voltage of Q12. Thus, the 5V power to operate the charging circuit is not held and charging is impossible. The charging circuit outputs CHG IRQ and CHG STATUS signals. When the circuit is supplied with 5V, the CHG IRQ signal becomes high level to indicate the instrument is placed in charge mode. The CHG STATUS signal is output from pin 13 (CHG) of IC18 to indicate the current charging status. If the battery is not connected, the CHG IRQ signal cannot become higher than 0.7V since D15 is electrified. Thus, the instrument is not placed in charge mode. The CHG signal is output in the following manner: CHG STATUS Output Level Low High Continuous Abnormal power voltage Continuous (rise) Start of charging 1 375s ± 0.225s Standby . 125ms ±20ms Continuous Charging 125ms ±20ms Charged 125ms ±20ms CHG IRQ Signal CHG swilch ON Cha/ping Mode 2 0s . ' * * CHG STATUS Signal in ordinary charging \ Start of Charging Mode Charolng In Progress Churging ConipleloiJ CHG STATUS Signal with some pending of charging 2 0s . Rending CHG STATUS Signal with abnormal battery voltage Charging in Progress No charging made CANT CHARGE" message appears aller 5 seconds Irom the start " . 2-18 CHAPTER 3 Troubleshooting 1 Introduction . 1 Internal Power Supplies 1 2 Faulty Power Supplies vs. Troubles Troubleshooting 2 1 All Specified Secondary Power Supplies Are Not Available in AC Operation 2 2 All Specified Secondary Power Supplies Are Not Available in Ni MH Battery Operation 2 3 +24V Is Not Output 2 4 + 10V Is Not Output 2 5 Vdd Is Not Output 2 6 Unisolated ±5Va Are Not Output 2 7 Isolated ±5V Are Not Output 2 8 Troubles Related to Charging 2 8 1 Charging Mode or "CHARGING" Status Is Not Effected. 2 8 2 Virtually No Charging Is Made Despite "CHARGING COMPLETE" on Display 2 9 LCD Operation Is Abnormal 2 10 No Key Operation Is Possible 2 11 Paper Is Not Driven or Abnormally Driven 2 12 No Mark Detection on Z-fold Paper Is Available 2 13 No Paper End Detection Is Available Or "PAPER END" Alarm Is Constantly Displayed 2 14 Program Contents Change 2 15 Buzzer Does Not Operate 2 16 No Waveform Recording Nor Alphanumeric Printout 1 . . 2 . 3-1 3-1 3-2 3-3 . 3-3 . . . . . . . . . . . . . . . 3-3 3-4 3-4 3-4 3-5 3-5 3-6 3-6 3-7 3-7 3-8 3-8 3-9 . . . 3-9 3-9 3-9 . 17 18 2 19 2 20 2 . 2 . . . Is Available 3-10 No DC Input Waveform Recording Is Available Any Waveform of All 12 Leads Is Recorded As Baseline Waveform of Some Specific Lead Is Recorded As Baseline... Checking AID Converter 3-10 3-10 3-11 3-11 1 Introduction . 11 . Internal Power Supplies First, understand internal power supplies and their functions. 10Vn The IOVn power is output from the secondary side of transformer. Since it is not regulated, it fluctuates depending on the load status and AC input voltage. It is transformed to various regulated voltages to be supplied to respective circuits. +10V Inputting 10Vn, the low-loss 4-terminal regulator outputs regulated 10V. However, if the input is lower than 10V, the output voltage will be lower by several ten millivolts than 10V. This voltage is supplied to the motor and the regulator which produces -i-5Vd. + 5Yp Inputting + 10V, the dropper type low-loss regulator outputs regulated -(-5Vd. This voltage is used as the power supply for digital circuits, such as CPU, of the unisolated section. + 24V Inputting 10Vn, the switching regulator outputs regulated +24V. A limiter is applied at approximately 0.4A referred to 10V, thereby lowering the output voltage. The +24V voltage is supplied to the thermal print head. In the case of FX-2111N, it is also used as a charging power supply. +5Va (unisolated) Inputting IOVn, the dropper type 3-terminal regulator outputs regulated -i-5Va. 5Va (unisolated) - Inputting +5Va, the inverting type switching regulator produces -5Va. ±5Va are used as analog power supplies for DC input, ECG circuit and AID converter. ± 5Vf (isolated) Receiving 10Vn through the isolation transformer which isolates the voltage from other circuits, the 3-terminal regulator outputs regulated ±5Vf. The ±5Vf are used as power supplies for digital circuits of the isolated section. ± 5Vb (isolated) ±5Vb are produced from ±5Vf through diodes. They are used as power supplies for analog circuits (input amplifiers) of the isolated section. 3-1 2 Faulty Power Supplies vs. Troubles 1 . 1 21 If 10Vn is not output... The FX-2111 does not operate at all since 5V, 24V and all other powers are not produced. . . 122 If 10V only is not output... The dropper type low-loss regulator cannot produce +5Vd, thereby disabling the CPU, etc. to operate. The FX-2111 does not apparently operate . . at all while 24V and others are available. 3 If +5Vd only is not output... The CPU and other components are not supplied with power. The FX-2111 does not apparently operate at all while powers other than +5Vd 12 . . are available. 12 . 4 If +24V only is not output... The LCD and keys operate normally but waveform recording and al- . phanumeric printout are not available at all. 12 . 5 If unisolated ±5Va are not output... . The AID converter and comparator do not operate normally, thereby making ECG waveform or DC input waveform a baseline with alphanumerics printed out normally. 6 If isolated ±5Vf and ±5Vb are not output... All 12 leads are displayed and recorded as baselines with alphanumerics printed out normally. 12 . . A Caution When checking the power supply and related circuits for troubleshooting, take sufficient care to avoid a short circuit. 3-2 2 2 Troubleshooting . 1 All Specified Secondary Power Supplies Are Not Available in AC Operation . (1) Check the FX-2111 is securely connected to the wall outlet using the specified power cord. (2) Check the AC inlet has two specified power fuses inserted and the fuses are not blown out. (3) Check AC voltage at the secondary side of power transformer. Disconnect the output jack, which comes from the transformer, from P1 connector and measure the voltage between pins 1 and 3 of the output jack. The AC voltage should be as follows: 9 9V to 12.5V for 115V version . 8 5V to 12.5V for 230V version . (4) Check supply voltages on PCB-6236. 1) Check voltages between the anode side of D2 and D3 and the cathode side of D1 and D4. They should be in a range of 10 to 16V DC. If DC power output is not available on the board while the transformer output is found normal, the contact of PI connector may be faulty or diodes D1 to D4 may be damaged. 2) Also verify that the voltage between the cathode side (TP22) of D5 and GND (TP21) is in a range of 10 to 16V DC. If the voltage between the cathode side of D5 and GND is not available, the fuse F1 on the board may be blown out or D5 may be damaged. A Caution When checking power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand. 2 2 All Specified Secondary Power Voltages Are Not Available in Ni-MH Battery Operation . (1) Open the rear cover of instrument and check the presence of battery. Then check that the battery is securely connected to the FX-2111 via the connector. (2) Check supply voltages on PCB-6236. 1) Verify that the battery voltage is available between pins 1 and 4 of P10 and between the plus side (TP25) and GND (TP21). If the battery voltage is not available there, the contact of P10 may be faulty or the fuse F2 may be blown out. 2) Press the ON key and verify that the battery voltage is available between the drain of Q4 (anode side of D6) and GND. If the battery voltage is not available there, the holding circuit around Q4, D7 and Q5 may be faulty or the connection of key panel may be inferior. 3) Verify that the battery voltage is available between the cathode side (TP22) of D6 and GND. If the battery voltage is not available there, D6 may be faulty. 3-3 2 . 3 +24V Is Not Output (1) Check the presence of IOVn (8V to 16V). Then verify that the voltage is aupplied to pins 10 and 13 of IC6. (2) Check operations of components around IC6. 1) Using the oscilloscope, verify that pin 4 of IC6 outputs a triangular wave at approximately 170kHz. If not, IC6 may be faulty. 2) Using the oscilloscope, verify that the gate of Q9 outputs a rectangular wave at approximately 170kHz. If not, IC6 may be faulty. 3) Using the oscilloscope, verify that the waveform at the drain of Q9 is switched at approximately 170kHz. If not, Q9 or D13 may be faulty. 4) If +24V is supplied to the thermal print head only, conduct the following: Press the START/STOP key to place the instrument in recording condition. Then verify that -I-24V is available at the drain of Q24 and pins 13, 14 and 15 of P5 connector. If the + 24V is not available there, consider Q24 or Q15 may be faulty or check whether or not signals are sent from the gate array (pin 21). Signal level of Q15's base input: High-24V ON Low- 24V OFF If +24V is available there, check the cable of thermal print head. 2 4 + 10V Is Not Output . (1) Check the presence of IOVn at pin 1 of IC22. To make 10V output available, the input 10Vn should be higher than 10V. If the input voltage is present but the output is not available, setting resistors R119 and R118 or IC22 may be faulty. 2 . 5 Vdd Is Not Output (1) Verify that 10V input is available at pin 1 of IC7. If the IOVn output from the secondary side of power transformer is lower than 10V, however, the 10V input at pin 1 of IC7 is lower by several ten millivolts than the specified -I-10V. If the input voltage is present but the output is not available, IC7 may be faulty. 3-4 2 6 Unisolated ±5Va Are Not Output . (1) Verify that the 10Vn input is available at pin 3 of ICQ. If not, both ±5Va should not be output. Conduct the steps described in 2.1 and 2.2. (2) Verify that +5Va is output at pin 1 of ICQ and pins 3 and 5 of IC10. If not, ICQ may be faulty. (3) Verify that -5Va is output at pin 4 of IC10 (TP2Q, TP53). If not, IC10, L2 or D14 may be faulty. 2 . 7 Isolated ± 5V Are Not Output (1) Verify that +10Vn (8V to 16V) is supplied to pins 6 and 7 of 12. If not, check whether or not 10Vn is output from the secondary side of power transformer. (2) Check the circuit around 12. 1) Verify that pin 11 (TP49) of IC20 inputs a 200kHz rectangular wave as clock signal. If not, IC16 or IC1Q on the oscillator circuit may be faulty. 2) Verify that pins 8 and 9 of IC20 output rectangular waves at 100kHz in reversed phases. If not, IC20 may be faulty. 3) Verify that pin 3 (TP60) of D29 and pin pin 3 (TP61) of D30, each at the isolated side, output rectangular waves at 100kHz. If not, check Q20, Q21 and 12, which may be faulty. 4) Verify that rectified and smoothed voltages in plus and minus directions are input into pin 3 of IC33 and pin 2 of IC34, respectively. If not, D29 or D30 may be faulty. If all the abovementioned inputs and outputs are found normal but isolated ±5V are not output, IC33 or IC34 may be faulty. 3-5 2 8 Troubles Related to Charging . 28 . 1 Charging Mode or "CHARGING" Status Is Not Effected . If charging mode is not effected, +5V (CHG IRQ) may not be output at pin 9 of IC18 or +5V may not be supplied to the charging circuit. Conduct steps (1) and (2)-1) below. If charging mode is effected but "CHARGING" status is not initiated, + 5V may be available on the charging circuit and at pin 9 (CHG IRQ) of IC18 but either the battery voltage or temperature may not satisfy the charging condition. Also, if +5V is not correctly output, the voltage may vary con- siderably and may be beyond the specified range. So conduct steps (1), (2)-1) and (2)-2) below. (1) Open the rear cover of instrument and check the presence of battery. Then check the battery is securely connected to the instrument via the connector and the battery cable is not disconnected. (2) Check the circuit around IC18. 1) Press the OFF key in AC operation and verify that pin 16 (TP44) of IC18 outputs + 5V. Also verify that pin 9 (CHG IRQ) of gate array IC8 outputs +5\/. If not, check the operation of + 5V power supply. . If +5V is output only when the OFF key is pressed Q12 may be faulty. . If +5V is not output at all even if the OFF key is pressed Q11, Q12, or D8 may be faulty. . If only +5V for CHG IRQ is not output check whether or not the collector (TP48) of Q8 is at low level. If it is low, Q8 may be faulty. 2) Verify that voltages between pin 6 (TS) and pin 9 (SNS) and between pin 7 (BAT) and pin 9 (SNS) of IC18 are in the specified ranges as , , , follows: Pin 6 (TP42) 1.01V to 2.0V Pin 7 (TP2) 1.0V to 1.78V . If the voltage at pin 7 (BAT) is beyond the specified range check the terminal voltage of battery. If the battery voltage is not higher than 8.0V or so, the FX-2111 is not placed in "CHARGING" status. Owing to this, supplemental charging is made in AC operation. Check whether or not 24V output is available at the collector of Q13 in AC operation. . If the voltage at pin 6 (TS) is beyond the specified range remove the battery and check the thermistor in the battery. Measure the resistance between pin 2 (blue) and pin 4 (black) of the collector. The resistance should be in a range of 3kft to 30kl2 (lOkft at 250C). Ambient temperatures should be proper. , , 3-6 28 . 2 Virtually No Charging Is Made Despite "CHARGING COMPLETE" on Display . (1) Check the circuit around IC18. 1) Verify that the source side of Q17 is supplied with +24V. If not, conduct the steps described in "8.3 +24V Is Not Output:' 2) Verify that pin 14 (MOD) of IC18 outputs a rectangular wave at approximately 140kHz; the drain (TP52) of Q17 outputs a rectangular wave at approximately 140kHz; and pin 9 (SNS) of IC18 outputs a triangular wave at approximately 140kHz. If not, check the following: . R18 and R82 including their mounting condition , . 017 Q10, 016, and D22 , . If above components are found normal IC18 may be faulty. NOTE: Charging current is detected as a voltage value at pin 9 (SNS), thereby switching the output of pin 14 (MOD). , (2) Check the number of charge-discharge times and the number of operating times after charging. Usually, the number of charge-discarge times is 200 max, though it may decrease depending on the operating environment. The larger the number of operating times after charging, the shorter the service life of battery becomes. Thus, if the charging circuit is found normal but virtually no charging is made despite "CHARGING COMPLETE" on display, the battery may exhaust the service life. 2 . 9 LCD Operation Is Abnormal (1) Check sure connection between the P2 connector on the main board and the connector of LCD. Check the connector terminals for any possible inferior soldering. (2) Verify the following voltages on the P2 connector: GND level at pin 1 (Vss) +5V at pin 2 (Vdd) + 0.238V at pin 3 (Vee) (3) Check the continuity of the following signal lines referring to "4.2 LCD Control Circuit!' Rs signal R/W E signal Signals of DB0 to DB7 NOTE: The LCD module used for FX-2111 has a built-in control driver and is directly accessed by the CPU, with MEMORY READY put. 3-7 2 10 . No Key Operation Is Possible (1) Verify sure connection between the P7 connector on the main board and the flexible cable of key panel. Check the P7 connector for any possible inferior soldering. Check the flexible cable of key panel for any possible disconnection and wear of the carbon part. (2) Verify that key signal lines except for ON and OFF keys are normally pulled up by checking TP120 to 125 and TP128. (3) Check the voltage at pin 20 (power supply) of IC15. Check signals at pin 1 (KEY) and pin 19 (AO) of IC15. Signal at pin 1 should be at low level at every one millisecond. Verify that pins 2 to 9 and 11 to 18 of IC15 are at low level when a key is pressed. NOTE: If only some specific key is not effective, the key may have a poor contact. A Caution The inserting part of key connector is made of carbon. Avoid frequent repeated disconnection and connection. 2 11 Paper Is Not Driven or Abnormally Driven . (1>/erify sure connection between the P4 connector on the main board and the motor cable. Check the P4 connector for any possible inferior soldering. Check the motor cable for any possible disconnection. (2Verify that pins 3 and 5 (TP103) of P4 connector are supplied with +5V. (3yerify that the following signal is available at pin 1 (TP101) of P4 connector. !< 5|19. 5;s ;< 5*L9. 5,s + 10V JUULJUU Paper speed 25mm/sec Paper speed 50mm/sec Check the signal at pin 14 (TP59) of IC25. (Refer to "2.4 Motor Control:") It should be 256Hz with a paper speed of 25mm/sec and 512Hz with a paper speed of 50mm/sec. Verify that the signal at pin 2 (JP5Q) of IC12 is at high level when the paper is driven. 3-8 2 12 No Mark Detection on Z-fold Paper Is Available . (1) Verify that "RECORDING MODE" is set to "MARK" in the program. (2) Check the mark sensor for any possible dirt deposit and inferior mounting condition. (3) Verify sure connection between the P6 connector on the main board and the flexible sensor cable. Check the P6 connector for any possible inferior soldering. Check the flexible sensor cable for any possible disconnection. NOTE: The mark sensor is the one and only part in the FX-2111, which is adjusted by a variable resistor. Before conducting the steps described above, therefore, check the variable resistor for any possible peeling-off of the paint lock and slippage. 2 13 . No Paper End Detection Is Available Or "PAPER END" Alarm Is Constantly Displayed (1) Check the mark sensor for any possible dirt deposit and inferior mounting condition. Verify that the magazine sensor switch is positioned in place and pressed when the magazine is closed. (2) Verify sure connection between the P6 connector on the main board and the flexible sensor cable. Check the P6 connector for any possible inferior soldering. Check the flexible sensor cable for any possible disconnection. NOTE: The mark sensor is the one and only part in the FX-2111, which is adjusted by a variable resistor. Before conducting the steps described above, therefore, check the variable resistor for any possible peeling-off of the paint lock and slippage. 2 14 . Program Contents Change (1) Verify that the lithium battery voltage is +2.5V. Check the lithium battery for any possible inferior soldering. (2) Verify that when the FX-2111 is turned off, pin 28 of IC2 is supplied with +3V from the lithium battery to back up the RAM. NOTE: Program contents are stored in the RAM IC2 and backed by the lithium battery when the FX-2111 is turned off. If program contents change despite sufficient lithium battery voltage, refer to 2 9 Memory Backup Circuit:' " . 2 15 . Buzzer Does Not Operate (1) Verify that pin 15 (or R73) of the gate array outputs a rectangular wave at 1kHz under the condition where the buzzer should sound. If the wave is normally output, R73, soldering of the buzzer or the buzzer itself may be faulty. In the case where no QRS-synchronized sound is generated, make sure that "QRS BEEP" is set to "ON" in the program. 3-9 2 16 . No Waveform Recording Nor Alphanumeric Printout Is Available (1) Verify sure connection between the P5 connector on the main board and the thermal print head cable. Verify sure connection between the thermal print head and the cable. Check the P5 connector for any possible inferior soldering. (2) Verify that +24V is available between pins 13 and 15 (TP112) of the P5 connector under the condition where a waveform should be recorded. Verify that pin 9 of P5 connector is supplied with +5V. (3) Verify that the TOFF signal at pin 21 of gate array or the base of Q15 is at high level under the condition where recording should be made. Verify that signals of LATCH (pin 2 of P5 or TP107), CLOCK (pin 3 of P5 or TP108), STROBE (pins 5, 6,and 7 of P5 or TP109) and DATA IN (pin 1 of p5 or TP106) are available at the timings as described "2.8 Thermal Print Head Control Circuit;' (4) To make sure of the correct operation of the RAM for thermal print head: . Verify that pin 28 of IC5 is supplied with +5V . Verify that JP3 is short-circuited by soldering . . NOTE: If electrical signals are normal mechanical failure may make waveform recording and alphanumeric printout impossible. So check the spring mounting condition and other mechanical status around the thermal print head. , 2 17 . No DC Input Waveform Recording Is Available (1) Make sure that "DC RECORDING" is set to "ON" in the program. (2) Check the P3 connector on the main board for any possible inferior soldering. (3) Verify that pin 4 of IC21 is supplied with +5V and pin 11, with -5V. If IC21 is not correctly supplied with power, the FX-2111 should not be able to record ECG waveform too. (4) Check the AID converter referring to "2.20 Checking AID Converter" If the AID converter is faulty, the FX-2111 should not be able to record ECG waveform too. 2 18 . Any Waveform of All 12 Leads Is Recorded As Baseline (1) Check isolated ±5V and unisolated ±5V referring to 2.6 and 2.7. Verify that power supply pins of IC12,13,17, 21, 31 and 35-45 (operational amplifier, multiplexor, etc.) are supplied with power. (2) To check the reset circuit and the control signal, verify that pin 13 of IC44 is not fixed to high level. 3-10 2 . 19 Waveform of Some Specific Lead Is Recorded As Baseline (1) Check the lead cable for any possible internal disconnection and inferior hookup. (2) In relation to the lead which results in a baseline, check the buffer IC, the resistance of resistance network, power supplies to lead selector ICs (IC36,38,41 and 45) and their connection status such as soldering. 2 20 Checking AID Converter . Check the AID converter if the trouble occurs despite the normal condition detected by taking steps described in Section 2.16, 2.17 or 2.18. (1) Verify that pin 7 (TP7) of IC13 outputs a triangular wave with an amplitude of ±3.2V and a pulse width of 520ns. If the triangular wave is normally output, proceed to step (4) below. (2) Verify that power supply pins of IC17,21 and 13 are supplied with power. Verify that the reference voltage of +2.5V is supplied. If not, check the power supply. (3) Verify that a rectangular wave ADCLK (L: 520(is, H: 480|js) of 1ms cycle is output at 1ms cycles from the collector terminal of Q1. (4) Verify that the low level S/H signal with a pulse width of 96ns is output at every 1ms from pin 6 of IC17, referring to 2 3 AID Converter Circuit!' " . (5) Check the output of comparator as follows. * Verify that pin 1 (ECG signal) of IC13, pin 8 (DC input signal) or pin 14 (battery/thermistor temperature) outputs a pulse which changes in the pulse width corresponding to the changing input signal. Verify that pins 4, 6 and 8 of IC12 output the abovementioned pulse signal of which the amplitude is limited to a range of 0 to 5V. 3-11 CHAPTER 4 Maintenance 1 . 2 . 3 . 4 . Replacing Power Fuses Replacing ROM Replacing Rechargeable Battery Operation Check by Self-testing 4 1 Self-testing Procedures 4 2 Examples of Test Results 4 3 System Errors Disassembling/Reassembling the FX-2111 5 1 Disassembling Procedure . . . 5 . . 4-1 4-1 4-2 4-3 4-3 4-4 4-5 4-6 4-7 1 . Replacing Power Fuses Power fuses are hardly blown out. But if they are blown out due to some reason, turn the FX-2111 off and disconnect the power cord from the wall outlet. Then pull fuse holders off the bottom panel as shown below. Insert spare fuses and put fuse holders in original positions. A Caution When replacing power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand. 2 . Replacing ROM (1) Remove the upper casing. (2) The ROM will be accessible on the PC board. Remove the ROM. (3) Mount the new ROM, taking care of the direction. The ROM's socket has 36 pins. If 1M-byte ROM is used, it will be mounted taking the full space for it. If the ROM is other than "IM-byte ROM, put it aside the rear of the socket (aside IC3). (4) Check the ROM is not contacted with feet of R18 when mounted. (5) If there is no problem, put the upper cover and fix it with screws. (6) Turn the FX-2111 on. Verify that the control program version on the standby display is the same as the label on the ROM. A Caution When replacing the ROM, be sure to turn the power off. Also, take care to install the ROM in correct position. tfC6 tft i) 4-1 3 . Replacing Rechargeable Battery (1) Turn the FX-2111 off and disconnect the power cord from the wall outlet. (2) Put the FX-2111 upside down. Loosen two screws off the battery cover. (3) Replace the battery with new one. Insert the connector of battery cable securely in place. (4) Put the battery cover in the original position and fix it with two screws. If the battery cover is not put squarely, rearrange the battery cable, etc. to neatly accommodate the battery in the room. (5) Press the ON key to verify that the FX-2111 is powered. (6) The new battery may have been stored for a long period of time and lowered in capacity due to self-discharge. So be sure to charge the battery after replacing. A Caution When replacing the battery, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand. c "v - V NOTE: It is recommended to charge the battery in the stock every six months. Also to make full use of the battery in the FX-2111, follow instructions in the operation manual. 4-2 4 . Operation Check by Self-testing The self-testing function permits you to perform the following tests: 1 Total Test 6. LCD Test . 2 . Sensitivity Test 7. Key Test 3 Time Constant Test 8. Status Test 4 Print Test 9. Automatic Power-off Test . . 5 Recorder Test . 41 . Self-testing Procedures (1) Press the MODE key to place the FX-2111 in the program mode. The display will be as follows: 1 AC FILTER (OFF) (2) Press the ImV (A) key. The following message will appear: MAINTENANCE PUSH (START) KEY ] (3) Press the START/STOP key. The display will be as follows: M-1 AC FILTER FREQUENCY (50HZ) (4) Press the ImV (A) key. The following message will appear: SELF TEST PUSH (START) KEY (5) Press the START/STOP key. The display will be as follows: 1 TOTAL TEST Press the START/STOP key once more. The total test will start to test sensitivity, time constant, print, recorder, and LCD. To proceed to the Key Test, Status Test or Auto Power-off Test, select it using the < or key. 9 AUTO POWER OFF 4-3 r ? 1/2 1 Z TEST 1 42 . . 2 Print Test ' n t a» « * 1 * ' : 42 . . 3 Recorder Test ; .... ;..> . Mm" RECORDER 'EST 42 . . 4 LCD Test 0 o 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 4-4 ! 4 3 System Errors . System errors are caused by troubles with hardware or software. Usually, they cannot be treated by users. If a system error occurs, the following alarm message blinks and the buzzer sounds: (Blinks) S Y T S E CODE M m R R 9 0 O R If a system error occurs during recording, the operator should stop the FX-2111 from recording (stop the motor). All the operation keys will be ineffective. Also, the normal operation will not be recovered unless the power is turned off once. Error codes and causes are as follows: Code 01 Type Description Overheat Thermal print head is abnormally heated to over 60oC. 02 ROM error Checksum error 03 RAM error RAM's read/write error 04 Unused interrupt 1 Unused software interrupt 05 Unused interrupt 2 Unused software interrupt 06 RTM error 1 Task No. to RTM macro is illegal. 07 RTM error 2 Task status to RTM macro is improper. 08 RTM error 3 Task's queue area overflows. RTM: Real-time monitor 4-5 5 . Disassembling/Reassembling the FX-2111 For maintenance, inspection and replacement of ROM and other components, you need to disassemble the instrument. Use the following steps to disassemble: The FX-2111 will be reassembled in the reverse order. Also, refer to the Chapter 9 Structural Diagrams" when disassembling. " When disasssembling or reassembling the FX-2111, take the following cautions: A Caution Disassembling/Reassembling Precautions . Be sure to disconnect the power cord and make sure the instrument is turned off before disasasembling or reassembling. . Do not remove the battery before disconnecting any PC board . . Take care that repeated disconnection of the key panel and sensor board may result in poor contact. . Use proper tools to loosen screws . . When reassembling make sure that all screws are securely tightened and all connectors are completely inserted. , PC Board Handling Precautions . PC boards are equipped with highly sensitive components to static electricity. . PC boards are highly sensitive electronic devices Put removed PC boards in a proper protective bag or take appropriate measures to protect them. . Handle PC boards carefully. A shock to them may damage the components. . Never insert a connector to the powered PC board nor remove the powered PC board. 4-6 51 . Diassemling Procedure Take the following steps to disassemble the FX-2111: (The procedure described here is for the model equipped with the battery.) Doublt Semus Screw (S) ~ M3x8 e7) , ~ g3 jo Step 1: Remove the cover of paper magazine and the paper shaft. Step 2: Open the battery cover on the bottom panel and remove the battery. Irom Ttanslormer 2-Double Semus Screw (S) , Irom RecodJer Motor Thermal Array Head Cable Sensor Cable Double Semus Sciew (S) . fe I M3x8 - 1/ 1 51 . * 0* 4-Double Semus Sciew (S) M3x8 . Step 3: Loosen four screws on the bottom panel and remove the upper case. Step 4: Diconnect connectors from the PC board and loosen two screws on the PC board. Then, the PC board can be removed. 4-7 Double Semus Screw (S), M3x8 3-Double Semus Screw (I). M3xB r - > 8 0 8 0 0 8 0 Step 5: Loosen three screws which fix the recorder onto the botton panel and remove the recorder. Step 6: Loosen the screw which fixes the power board PCB-6238 and the transformer onto the bottom panel. Disconnect the connector from the transformer and remove the power board. 2- Double Semus Screw (S) , M3»8 m 8 8 el 0 5 Step 7: Loosen the screws whch fix the transformer onto the bottom panel. Then, remove the transformer and grounding parts. Now, disassembling is complete. 4-8 CHAPTER 5 Periodical Inspection 1 . Periodical Inspection 1 1 Checking lead cable and power cord 1 2 Visual inspection 1 3 Cleaning the FX-2111 5-1 5-1 5-1 5-1 1 4 Self-test and total function check 5-1 15 Battery check Safety Inspection 2 1 Measuring leakage current 2 2 Measuring protective grounding resistance 5-2 5-2 5-2 5-5 2 3 Remarks 5-6 . . . . . 2 . . . . 1 Periodical Inspection . In this chapter, we describe the inspection to prevent troubles and let the FX-2111 keep the sufficient safety and complete operating condition. Perform the following at least once a year: Check the lead cable, power cord, and grounding wire for any possible damage. . Conduct visual inspection . . Clean the FX-2111 . . Perform the self-test and total function check . . Check the battery . Check the leakage current Measure the protective grounding resistance. . . As with simple inspection, recommend the user to make it every day or week. But let the user refer a complete function check and troubleshooting to the service engineer authorized by Fukuda Denshi. 11 . Checking lead cable and power cord Visually check the lead cable and power cord for any possible damage. Check also the connectors for any possible looseness. If a damage is found, replace them as required. Merely applying a tape may not repair the cable sufficiently. Also measure the resistance of each electrode to check for internal cable disconnection and damage. 1 2 Visual inspection . . . . Check that mechanical parts such as screws are securely fitted Check that connectors inside the FX-2111 are securely connected Check that there is no trace of damage If a loose part is found tighten it securely. . 1 . . . , 3 Cleaning the FX-2111 . The recommended number of cleaning times depends on the operating frequency and environment. For cleaning, wipe the casing with a properly wetted soft cloth. You can use the cleanser for tableware by weakening it with water. However, take care that the cleaning solution or water may not enter the inside through openings. Finally, wipe off wetty substance with a dry cloth. 1 4 Self-test and total function check . Perform the self-test described in "4. Maintenance" and check the total function of the FX-2111 to ensure that the equipment operates safely and completely. 5-1 15 . Battery check Measure the voltage between plus and minus sides of battery to check the capacity. If it is below 9V, charge the battery. Also, if the battery has not been charged for six months, charge it. If the battery will not be used for a long period of time, remove it from the FX-2111 and store in a cool place. If the battery is below 6V, it may be difficult or impossible to charge. In such a case, replacement is required. 2 . Safety Inspection To ensure the safety of FX-2111, it is recommended to perform safety inspection. The test methods and measuring instruments are stipulated in the standard for safety test. It is considered extremely difficult to fully follow the standard for the safety test at the health care site. In checking for maintenance and inspection, therefore, measure each specified value as a rough rule of thumb. We describe here examples of simplified measurement of leakage current and protective grounding resistance. A Caution If you find a value which exceeds the allowable level, be sure to let the user avoid using the FX-2111. If the user operates the FX-2111 as it is, he/she may receive an hazardous accident. 2 . 1 Measuring leakage current Prepare an instrument to measure the leakage current (electronic or digital voltmeter), impedance device and power switch box. The impedance device has the following configuration: R1: 10kn ±5% R2: 1kfi ±1% R 1 C1: 0.015tiF ±5% V: R 2 C 1 -r Voltmeter V You can obtain a leakage current by measuring voltages at both ends of the impedance device. 5-2 The digital voltmeter shall indicate a true root mean square value to a composite waveform of a frequency band from DC to 1MHz. If such a voltmeter is not available in hand, you may use a commercially available digital voltmeter for the purpose of simple maintenance and inspection. However, make sure of the frequency band of the voltmeter and note that the voltmeter will not indicate a leakage current at a frequency exceeding the capacity. (1) Example of checking ground leakage current (current flowing in protective grounding conductor) Power Switch Box Polarity Switches oa POWER - - ob - Wall Outlet O Medical Electronic Instalment a °b - GND GND - o Oft ' A Point 8 Point Power Cord M D MD=lmpedance Measurement Device See the figure above. A gounding terminal and the power switch box which allows you to switch the polarity facilitate measurement of ground leakage current. The procedure to measure ground leakage current using the power switch box is as follows: . Measurement of ground leakage current is made by measuring voltages at B point (grounding conductor of power cord) and A point (wall grounding terminal) under normal condition and single fault condition. Normal condition . Measurement of ground leakage current under normal condition is made by switching the polarity switches to a" positions, then to b" positions. Thus, measurement is made in two ways. " " Single fault condition . Measurement of ground leakage current under single fault condition is made with a power fuse removed by switching the polarity switches as above. Since the same measurement is made with another fuse removed, measurement for single fault condition is made in four ways. 5-3 (2) Example of measuring enclosure leakage current (current flowing from enclosure to the ground) Power Swdch Box Polanty Switches ; POWER ! Oj- - - ob Medical Electronic Insirument Wall OuUel GND GND T-O A Point B Point Opsn lor ilngla lauli condillocyPower Cord | M D \- Metal Foil Probe MO Impedance Measuiement Device Or Apply a metal foil of 20 x 10 cm to the isolated enclosure. Put the probe to the metal foil for measurement. Measurement of enclosure leakage current is made by measuring voltages at A point (wall grounding terminal) and B point (desired part of enclosure). If the enclosure is isolated, stick a metal foil (e.g. aluminium foil) onto the enclosure with a conductive tape and apply the probe to that metal foil. Measurement of enclosure leakage current under normal condition is made by switching the polarity switches to "a" positions, then to b" positions. A value measured under normal condition will be a very small value near zero. Measurement of enclosure leakage current under single fault condition is made with the wall grounding terminal and the grounding conductor of power cord removed and by switching the polarity of " power source. In leakage current measurement, take a maximum value as the measurement result. The following are maximum allowable ground leakage current and enclosure leakage current: In inspection, it is important to compare measurement values with past ones to check for any outstanding difference. Maximum allowable leakage currents (117/253V AC) Ground leakage current Enclosure leakage current Normal condition Single fault condition 0 5mA 1 0mA 0 1mA 0 5mA . . . . 5-4 22 . Measuring protective grounding resistance According to the standard, protective grounding resistance shall be measured by letting the testing transformer flow an AC current of 10-20A from a power source of which the voltage with no load does not exceed 6V. Practically, however, it is not easy to prepare such a testing device. We show therefore a simplified method as follows: V O o Power Iranslormer shall be able lo let a curfent of 13A How at a voltage ol less than 6V with no load . Medical Eqjipmenl VW A Point Power 30 Transformer Or C Point EO (SOW) 6 A C 6V Metal Part VI (Voltage between A and C) o 2 30 V V2 (Voltage between B and C) R: Protective Grounding Resistance B Point . O t El Power Cord ' Simplified Measurement of Protective Grounding Resistance The testing circuit above allows you to measure the protective grounding resistance of equipment using an AC current of 2A or so. Connect the power cord of equipment to El and the metal part (such as equipotentialization terminal) to EO. Make the conductor to EO shortest possible to suppress the contact resistance. Flow a current of 2A or so to the equipment and measure voltage V1 between A and C points and V2 between B and C points. Since the flowing current is the same, the subject protective grounding resistance can be obtained through the relations between the voltage ratio and resistance ratio as follows: R/3 = V2/V1 " . . R (protective grounding resistance) = 3 V2A/1 (fi) Measurement current may be 1A or so. With the standard, the resistance between an accessible conductive part and the protective grounding terminal shall be 0.1 Q maximum. In measurement at the grounding conductor of power cord with the power cord connected, the grounding resistance including that of the protective grounding conductor of power cord shall be 0.2ft maximum. Protective grounding resistance Resistance between protective grounding terminal and potential equalization terminal 0 10 maximum Protective grounding conductor of power cord and potential equalization terminal 0 20 maximum 5-5 . . 2 3 Remarks . Abovementioned measurement of protective grounding resistance requires the power transformer which can flow an AC current of several amperes and a testing fixture. If you cannot prepare such a testing power transformer, you may measure it with the following method to obtain a rough rule of thumb: Measure the resistance between an accessible conductive part and the protective grounding terminal to verify that it is lower than 0.1ft. If you measure the protective grounding resistance with the grounding conductor of power cord as the protective grounding terminal, the resistance shall be lower than 0.2ft including that of the protective grounding conductor. You need also to adjust the resistance of the testing lead or probe used for measurement. Medical Ecjjipmenl Reslsianca Measuring InsUument Testing Lead Accessible Conductive Part < 0 . (e.g. potential equalization terminal) 1 fi Protective Grounding Terminal Potential Equalization Terminal po-m nvj Protective Grounding Terminal Simplified Method of Resistance Measurement This method considerably differs from that of the standard. Consider the result as a rough rule of thumb. Protective grounding resistance Resistance between protective grounding terminal and potential equalization terminal 0 1ft maximum Protective grounding conductor of power cord and potential equalization terminal 0 2ft maximum 5-6 . . CHECK SHEET FOR PERIODICAL INSPECTION Inspection Date: Inspected by: Model Number Installation Site Serial Number Years of Use Check Items Judgment/Measurement : Remarks (Repair Required) VISUAL INSPECTION 1 Cords and Cables . 2 . Power cord and lead cable for damage Pass/Fail Connections of power cord and lead cable Pass/Fail Casing and Accessories . Casing for damage such as crack . Casing for dirt Pass/Fail . Labels for fading letters Pass/Fail ' Pass/Fail . Screws for loooseness and dropout Pass/Fail . Electrodes for dirt and rust Pass/Fail . Expendables and operation manual Pass/Fail 3 Connectors . Pass/Fail . Power and lead connectors for damage and looseness r - 4 Others . . Power fuses for breakdown Pass/Fail Protective grounding terminal for damage Pass/Fail Switch panel for damage and dirt Pass/Fail Thermal print head for damage and dirt Pass/Fail Paper magazine for damage and dirt Pass/Fail LCD module for damage and dirt Pass/Fail ELECTRICAL INSPECTION Pass/Fail . AC operation and battery operation . Time constant test Pass/Fail . Recorder test (paper speed & sensitivity) Pass/Fail . Printout test (for missed dots) Pass/Fail . Key test Pass/Fail . LCD test Pass/Fail . Auto power-off test Pass/Fail V) . Battery charge Pass/Fail SAFETY |jA Pass/Fail . Ground leakage current (normal condition) pA Pass/Fail (single fault condition) |iA Pass/Fail . Enclosure leakage current (normal condition) (single fault condition) |jA Pass/Fail Q Pass/Fail . Protective grounding resistance OTHERS 5-7 Charge OK/NG CHAPTER 6 Circuit Diagrams Overall Block Diagram Overall Block Diagram Main Board Circuit Diagram, 1/9 to 9/9 Power Board Circuit Diagram 433-5125 6-1 433-4893 6-3 523-2020... 6-5 to 6-21 524-2021 6-23 Sensor Board Circuit Diagram Motor Module Circuit Diagram 524-1959 524-1960 6-24 6-25 Battery Pack Circuit Diagram Overall Connection Diagram Overall Connection Diagram LCD Connection Diagram Motor Connection Diagram 524-1961 433-5126 433-4894 434-4895 434-4896 6-26 6-27 6-29 6-31 6-32 r . . . DC Input * Main Board Pr.B-6409 Bult.r Amplifier' , i Power Supply ' : i A C 1 1 5 V > I i 5V - i 5 u I Isolated Section rs c Butttr Amplifier Preamplilici ConsUnt/ it Reset ' Circuit u n ModulaiJon Middle Ampliner i25 I 5 . Tranjlormer Circuit ! fi! 1 - Time Resislanct NelworK lor Lead S«lectior Demodulation Circuit I-> I . _ Amplifier 1 Control i Pustv Power Circuit Isolation Translormer i . Pull ' 21 I . Detector Detector Circuit Circuit PCB-€410 i: i . - - - I . I * i . Digital Power + 5V i Rectifier Circuit Overinpul AC Inlet Board \\ \\ ' Supplies Oscillator R-wav* I i ' l I Circuit 1 . ' J +VFB I I : i +5V j- Generator ~ +5V. I . , AC Input i i US Triangular |_ . "" i ' < £ IA Amplilicr r-»Wav* liaV T J r- f jlse t Generator Circuit i T Trans I or mc/ : i ; i i + 10'/ i , Recorder Photo Photo Coupler Coupler i 0 t A Sensor Board: - Sensor Circuit . ' PCB-62 39 Gale Array Motor Control R ROM E AM27C512 S 54Kbyte QS6-F L - . - - . .1 0 Charging Circuit »| A V A T CLK SRH2254-12 16HHz 8Kbyte lor Thermal Prinl Head I RAM SRM2264-12 i i _ -» RAM . i S S T i i i U t ' i B TEMP Prinl Head + 24V s 4 \ Thermal ?. A U i 2HHz In B Motor Module _ H063B09 T FD88O07-AC NF-26P1/25|* KM ; i 1 i CPU 7t , i , I :i :i ji r SKbvte Ni-MH r LCD NOH202AOO Battery PacK i Key Panel i DRAWING NO. TITLE Overall Block Diagram MODEL NO. ASSEMBLY NO. 433-5125 DATE 96.10.9 FX-2111 (USA) 6-1 , DC Inpul . 3 - Main Board . pcb-6409 Butter ' -.1 Power Supply 1 ; Amplitief - AC sv i o |i Isolated Section rs i c Time Resistance 5 Network lor Lead Seltctior o V Preamplilicr - xB Reset ' 3 Modulation Amplifier Circuit 125 Circuit o 1 1 Middle ConslanU - Amplilier Isolation Transformer' Demodulation Circuit Amplifier o T 2S1 ! 5£ xl.25 Triangular I r-SWave I ItnV i Power Circuit 1 , u Isoiation k Translormer - PushPull Circuit in AT Digital Power i + 5V * i i Supplies Oscillator Rectifier Circuit Oelector Circuit Detector Circuit 5 - Pulse Generator Circuit t I Photo Photo Coupler Coupler CPU HD63809 2MHz D Recorder FET A T Sensoi Circuit . PCB-5239 X - Gate Array 5 Motor Control NF-26P1/25 < R ROM AM27C512 8 S 64Kbyte ? i R AM CLK SRH2264-12 16MHz 8Kbyte i U S Charging Circuit I > AVJT i S i I > - TEMP Thermal Print Head 056-F +2 V i D E A FD88007-AC Motor Module Transformer + 10V v Sensor Board: pcB-e4io 11. _ I iVFB i Overinput AC Input Generator i5V Amplilier Control R-wave , AC Inlet Board I 3 . + 5V fc U 230 V B RAM U SRM2264-12 S 8Kbyte J . r Ni-MH LC 0 Battery i Pack Key Panel NDM202AOO - i lor Thermal Prim Head t DRAWING NO. TITLE Overall Block Diagram MODEL NO. ASSEMBLY NO. 4334893 DATE 96.10.9 FX-2111 (CE) 6-3 noi!) i miss omtnsi smcifih . . AU IEC1ST01S tit IN Q ±\%, I. || , ALL cmcims tit IK U F VBF-f VRF - uPC4064G2 IC37 1 . 4 4 _ 1039 <D tfl. tin IIS! PAH 4 . IC31 uPC4054G 4 14 IC35 TC-4051BF IC3Q TC-4051BF is 16 TC-4S3flRF TC-40S1BF IC44 SPICUl SYHtOlS UUCE - L . JUL 7 7,a 7,ft 16 A 16 3 7,6 16 7 ,a VBF- I1 < C97 Jcxot, )<: Ci 19 Cl IB ±2. T TC-40g4BF IC451TC-4051BF <M iM 1!S1 PtD T T T FC Fl TCI? .TC-4Sfl4RF F2 7 +RV - -RV - VDD -14 4 a IC13 uPC4074C IC15 TC7W04F 4 TC17 TC-40S3RF a - 11 - ICig TC-4Ba4BF 7 - - IC20 TC-74PC74F 7 - - - u PC40B4G HEF4046BT 8 TC7W02F 4 4 IC1 TC2 QM27CS12-150DC SRM2264-12 IC3 ICS. HD53B09E ICS FDa6007-PiC 5RM2264-12 . C93 jcil? |ci25 |ci28 14 T L4_ T C94 \ci Ifc |ci2t ]ct.27 T T T T T J 4 IC26 IC27 - 6 JL4. _ IC21 SVF .SWF IB 9 FC FC E 1 VDD VDB 14 25 - 14 - 26 6 VDB 1 7 14 IB 40 65 ' C I I Ci \ CC |ci3 jcT« |cZ2 ] C30 j C31 | C 4> JTTTTTTTT 26 3 26 63 j | C3I joi Jclfc | C37 | COS jcfcl lct.6 ] C68 J_C7 3 jJTTTTTTTTT 90 TC11 TC74HC24BftF IC14 TC74HC04nF TC74HCB41ftF IC15 IC23 TC74HC04RF 1 7 2 V J 4 _ 1A 22l ±4. TITLE DRAWING NO. Main Board Circuit Diagram, 1/9 MODEL NO. FX-2111 ASSEMBLY NO. 523-2020 DATE PCB-6409 6-5 95.01.09 co1 »H3 o o - - o o - < 2 OK s mil r i 3 . IIS 2 OK JL ao rHI-- «i n 163 12 5 4 It IC37B . 11 X4 > %7 " > rnuj n tai ni&i caj 20K 3K 0 -VW-t . 21 1 'C 19» «l»T tlK IJ IQH >-fl. ,_ 0 1 -L. > | C 10] uPCIC 20K iC3a,.? f 2,OK - ICS 10 rc P««' 1 3 5 20K 2 5=| | LI LU. I t11170 > 1- 100 WPC4044 434 | 3.5n > J* 1 SS14 Jl J T rc >] > 30QK LSS302 uPC40t* If <( a? 20K » fC ZuK «1 . 201 1 10M n i hi H 183 t 40K 20K » 1 7 I - 5VF . IK I NH 2r 032a T 1SS301 _ . C 104 HQ* ' 1 104 S iBsr~ > «2 If > JOK 522 <.i/jsv xo . «J [CI 13 4 I JL <5± f rc t NH . - ' UPC4044 1 1051 40 X HOP tg41 , ?- ai »4 R20« 30K XC - sm i- 02 O %- QJ CLK [ j 04 oc t 3: Cl 12 0 I cue . 220 3t LOK UPC40ft4 I OK . U .... I NH 1035 IK 0223 Cl IS S IK 0 2 4 C TOTi [ C 120 ClOl I 3io0i.r 4 0?S I « I 12 Cl 14 4 '~ TWPPrtT A l A2ao LOK IK Paga svr c 1cj»o Q4Q runs 4 oo**r 4093 Jl mi' Pa«« UIIO04CSB-1.2 0 9 HOP 131 Xft2i4 ? 9. IK . 03 33 UN«3 If uf*C4044 11214 > 5 IK ? A ,c" It 1 > 113 19 C I rc 4 . B 10* ? i «i u JK 13K0 ] .4KB rnua T 3 t 1*1 22 IC370 n ia2 JL 40 X I tC3»« T 01S| . . 294 017 CB7 2 -"vw '0" . 210 I) HOP <1K [ c04 X X C04 140K < , . VW . IBM 7 OK IC3JC "" i- JJ . 1« rc:200H> 0.0013 0 Aiit IC3T0 XL f aoac-isscicr > I A JJl 0 01 I -svr e . Ton- V 1S0P 4093 Q CO Li 44 1 IC400 C42C «»41 <1K C I - lOK F _ » . 342 * 1C430 vr rC40«4 00. UX5 212 *IK V L«0H 1C43C 044 - rnM2 rs "140 svr JL Q Li 11 2 C7B LirC40«4 SIM o.oca [ C 123 1S0P . 1 BO 941 rnu2 . 174 . It 7 -S\J~V- 1>- IK i J 1 OK JL p 9 7 « 17* 10K PC rc40*4 10K in -T-n <.pc404 4 o.jj ; i?i 5bi7S /300K } 140K 1 pe»o44 10K rc rc <fl 1 73 II . 24S 5 - . .. A p.« - 1 04 3 UN 112 R210 uPC33 5 _ _ 24 oloaii HI- l 30 2 * 2 PC33» _ e . IK _ zook 4 n2:e C122 »2S2 : IJ IC43n IK 2 2 OOP . 24* Ji --*/W- O . 4 t I C4 J» SiOP . 2«0 »24» rc R 2SS 3or 0 . ? . IK C 134 C 1 32 - A224 49aK C I I 1 - fll4ft C133 6ok PC40%4 5 1124 1 ' 1C43U 3 .003 3 . . svr svr * /W HI- 0 22 rc rc ' CI3S 2 IC42n " - iC3in ?wr . 234 200K 1 . 230 . R73 . C79 rc R7S1 . 169 0 22 5 . IK M23T uPH131 ic«2o 0234 7 } JJL .301 ,J lOM 4S30 . 23ar 13K fl 2 3 Z JL " CO IC4aA 4S3a «? 3 f /tiOK . 247 - IOK S IK . 244 . CI g) 1-0 uFC 4064 . 25V| C131 HO' DRAWING NO. TITLE Main Board Circuit Diagram, 2/9 MODEL NO. FX-2111 523-2020 DATE ASSEMBLY NO. PCB-6409 6-7 95.01.08 R12(. Past « N 1 7 B n bfi- 2K v/V > Ri 28 > - <1I> I- 2K C47 K 1 79 1' SI* 1« C22 T 1 5514793 j . zr. 2 30Kr 120KF IO00P Rlt6 rtu nBD p 10 ' 5511391 B 0 Pa?? f c : 25 > 03 2 age 0.022 2 C21D E j > R4 a Li 1M 1O0OP R1CS 1- r 1M k 1 25 D O1 . 1 n UPC«b6«(jt) D . 0 J 0.01 IT 3 L3 X 0 z uPC«07a * 1 63 1 02 5S1«393 T 10O SSI 4393 . 3 2 FC Irc«5e« . 1 ? ? 0 11B ISS302 V O 022 . E J 5 9 zo R56 c 1 6 1 1 in Kb 2 I NH 1 C 13 C 2K R57 .10 2 L 00 E2 D 19B 1SS302 ~ 3 . P»9* 6 I" R 1 IS UPC4064 1 B2K 1SS302 R«5 RSO I C 130 . S-C8036-01 rc«58« D20e . SSI 4393 IK 1 0 19 3 Rill at v 0 02 2 K 7B 3 vv, fa C J 2 R 109 R 106 Pasff E 1 I C2 lO P3 20K ] C3e _ rc jjc E uPC407a 5V n v 20K T C23 _ IC 128 R43 7 iOOOP 5 J C 13« 2K 0 in OOP 5 ce5. 0 J C5B I w CS7 12 ce7 in CfcO C56 Dl 90 ZK "ess 1SS3D2 1DO z ICIJO R4£ 5 IP v»r-rr (2 . 5V > II SV - C5 4 E 1 2 C2 IB 78.7KB R5« R4B lOOKB J0D T 330OP R 114 6 R53 PC4074 JL. 1-I 20K 1 «' 1SS302 022 1 C 13B I C2 1 C 39KB 7 (EaIIi£Z> e 10 n Rl 13 UPC4064 7F.7KB R5 1 Rl 12 UPC4064 UPC4074 30K t-3 76.7KB C55 2 - *Z .2V J Cfe.25 V LSB) 22 S60P . El 4. E 1 C 1 P»gr t C45e« D 12B Z 1 . Z R130 P«ar 3 R 1D3 B Pave 6 018 5 . IK UNS112 R 1 OB B2K < t enp ysm 1 1 p,«r 6 I cizc 5/M P»9r £ " C45HC nToi DRAWING NO. TITLE Main Board Circuit Diagram, 3/9 MODEL NO. FX-2111 VOL > 523-2020 DATE ASSEMBLY NO. PCB-6409 6-9 95.01.08 - <-i . 6vn I Pa9,' 7 -E13Z> p,*. 3 -(Tfti- TP60 <Tftf 1<rBFi P»aF 3 Page 6 Page £ TPS 1 gg0 ~ IC33 Page > ~ 3 030 Njn78L0SUa TP69 . . 3 yj i" rl 42 1SS302 :T2 1 55302 <n2!D- ySKT.T £. VDD 4 2 . R123 e WV- - C 107 . CiOl 22Z 2 s 0 . 1 T P7C 0 13K - Ci02 <zsso- 2 f " a f *3 2 ci3e 33QP _ 4 2P I5_ I €34 1SS302 HJI"179l 0SUn . Q E2 ? R 12 1 _ 7P71 921 P R D T P4 9 CLK<- 1 RL20 I 13K < 5 029 3 3/2S<y 11 L U7'»«>C74 3 430 D23B 3 Rioa D33C ' 'W - T JSKA- inrn ao - T o IC20B H1.31 155302 42 .2 1SS302 C93 Pfi'se 2 3C37,39.02 Pauer T " Pin (?) | TP72 R10S R79 W** - T IK 034A 'm; 1SS302 IC16« IC16B TC7W04F TC7W04F T "vV 2 . 2K Cqfe 100C", 1000» IC16C IClt S IC19C C10O T O. 1 4584 2O0KH2 FQ C74 I C29 PS26S2L I NC Pag« TC7W04F 6 1| O . 04 7 0152 2 430 TPS ICi90 C72 tC28 NC P jge <ftt1PDaTO fS26S2L | 6 _ 2 1| 0 047 . . 5 V- R149 430 - 7P9 ic j.9e 4 - Past IC30 5VF PS26S2L 2 . 6 NC 1 [hp~ 3 TPli IC1.9C R222 TC4SB4 R102 3K 15K E2 DRAWING NO. TITLE Main Board Circuit Diagram, 4/9 MODEL NO. FX-2111 523-2020 DATE ASSEMBLY NO. PCB-6409 6-11 95.01.08 V/OD 6 Rise 20K P6 08370061000800 CKYOCERO ELCO I C 12F R 159 6 13 I 2K r 04584 C77 To Senior PCe-6239 E2 2 R160 % IK 7 C76 0 3O I €270 ft" a . I C27B N. 1 , 02 0270 R ISO R iS6 IC19D IK 4584 k L3k t . ozeo IK 155 ISS302 UNS212 1S5302 IC7U02F S2K IC7W02F C7 1 Q .022 R 15 X C 75 240K 7 C2 0 E2 E2 Page . i R1S7 C2 9 7 C2 C2 0 1 R 153 ISO 6 R154 R 132 ISO - 929 /N/V P4 1 K iC2oa HI 33 2 2SA14e3 027 JN5212 S as > C2 >CLK cJ>-i T a L Y 74«C7<I 2 250 1SS302 3 E2 HZ? 7 3 baov GEEEE UM51 12 V0C 1 PA9e IK 026 UNS212 Ra9« 2 Dr3A-6P-20SO 925 > CI N PP VCOUT p i SIN Pas* NC R 145 CX VCOIN [MH 16tC R 1 R2 OCnO ZEN I N 1 - J Rl 40 I Z 10NC D T C CG5 PC35B Lj C70 4 7 X2SV . R14*. L2 HCFOUI QUICNI L2 cc L2 R T 7 0 R137 LS L2 C i 20K C64 4 . T/SSV T 24K R 138 24K R 135 I 4-3K . 7 C2 i-«<> C2 C2 R13« 0 5 1./-0. 2S J . CNO 5 1SS302 IN2- R 1 42 lOOK NCTT 2 a 14 7 10 OK 2 8148 IC260 D26A IN2R 1 41 & I OK I C26B P2 2NC U - c i a i C 1 C62 . . 2 ? S 228 UN51 12 . uPC 4 9 4 C5 C69 2200» R 1 43 5 IK 7K 7 C2 C2 ce3 R13G 0 . 1 1K R13? 3&K 1 C 1 2A i K 1 44 1 K 1 t <i5b4 DRAWING NO. TITLE Main Board Circuit Diagram, 5/9 MODEL NO. FX-2111 ASSEMBLY NO. 523-2020 DATE PCB-6409 6-13 95.01.08 2SC R 14 22; Jfcv 2i 6C 2 s P5t S 55 I! r, " A as | O NWI 5-a n ib« -a r mo ei i - 11 1 AC AC it vis C 1 ! Si H-< 1 "1 A? il... P*9# 17 ;3 9 B9< 26 ST RST ' ST OF Y2 - - VOC C2 V7 < KE Yd 2 OI127C512-1S6DC tic "D* 1 «t. E2 D7 Old DC D93 fid 2! SI It- e« * R92-Bfc iJ 6iC NSt it at . NOTE J B9- 0D£ - 12 2 » IDC < CMC - 5 T O T u£ 20K 2QK F*q. " MS g.«..m«l!5iii MSfEE rifM ass -e£a jfrlff 3 ft? IE si etT vol BIO 1 240k Thro 12 - TmmDO -S-5 -" P i xo? cS: C5 f 10 t A 1 1 aii CSI we ft 12 oE 9 flTTu 1 [CI 1 i JF3 N <4 8gi ; s-§11 ? DE7 2 rH 01 R > 1* TMno= -S-i LOC h 01 V ISSI il I C 1 19 f P30 S < AQCLK Fa** S Fag* 3 NOT I OK OMFD sHhiigfsJiI ill NDFCISSKZR 0 12 5 ES S It23F "":C23C I C23D C2 IlOTCH IC2SO i [SJ2__CiiN> r > R251 2 _ W0TC2 . NO 1 SI - _ R01 in U*« U*> No. C 3:S 12Kbjt <F_orr ici-ac f c Ktei 1 RAW 1t-.«n connect JF1, JP3 256Kb.it RAM t *n connrct JF2.JF4 < 3: Jlb it 5 r - {> fe CHC JKC__> - JPfc - R7? 9 i 1K»r> Us* Soc-t connect JPfc infen co*-.n*c-t JPfc. JPB DRAWING NO. TITLE Main Board Circuit Diagram, 6/9 MODEL NO. FX-2111 523-2020 DATE ASSEMBLY NO. PCB-6409 6-15 95.01.08 6 01 -* 03 nC PCTCCT rpBBOO? " 1 3 " f D3 3 - . l a IP M T f R T> XT 0 @ 0 »3 2* OS D3f S«i» CND 024 2S J O 1 C 12 2 R.* 2900 BL. 0 1 BM 1 -Ofe 2 R3 C16 I T 09 i'3 lit. 2DOM < mo ' S- > 20K 4 AH R«2 3 4BK UM5212 . o-® S NMRCN/NM 0 C T T - C * 0 ® 1- c 08 Tf>«8 0 ON CI37 086 3? M * - >J«W o IS 272 IOO/2?v UM5ti2 B ( *2*v J . 3 I OOP C9 ® > i [C22 ® 2 Ion .i > I i D 1 0 013 . 1 « /3SV . /1*V ® BJ IB UN5i12 > C17 4* ik r 9 11 25«1 2 @ 1553D1 C29 "t? 1 20K 0 . No t kl IPS? J OuM C 19 W8® L ISS3D2 UN3212 > Q 17 ZSJ2«>SS Di6a D21 D.DJ vi n 3L -wout oen ) z f «70uM KB J 100 ® I o- i »a4 ® v TP2a <n > n 7 2 R71 Dirs * L 021 j k? uiza 12 c 12t( IP3 Rt2 3 . 4 ex OVEN % BID MOO 2SK T Ml S R63 ® C? E2«V * S 1 lODw C3 VS5 I T c« i lOD 6q2oo3 ® 0-15S30J D-1SS102 *JK7aLOSU* NJMT9LOSUA * 2SJ24S T P45 f f 1 I T ) OS-CON ?ft » 1 - out OUf I i 2 - GWD 2 - IN 1 ~ CA'C 3 - In 3 - COW COHBON Out 2 3 - [>R*IN SOURCE « - DRA1n DRAWING NO. TITLE Main Board Circuit Diagram, 7/9 MODEL NO. FX-2111 523-2020 DATE ASSEMBLY NO. PCB-6409 6-17 95.01.08 P7 0637015 1000600 CKY0CERO ELC0> nop E - o SCNSI - a o - i 1 2 2 3 3 4 4 s s i 6 7 7 e G 9 9 10 1 0 11 1 1 12 12 o o - o - NC SOU - - SZL NC (key? > 4XEY3 ~> fKE Y4 > - _ ST CRT /STOP O - O- 1 mV O - - O- INST O - O O ON - O - O - P»9* 7 P*9 c & (KEYS - (OF F /CMC 1 > to - _> > _ 13 L 3 11 11 IS IS 0 o fc - O- OFF/CHG - - P * S r {X E Yfc ~ > NC NC © E2 rxr2A.U..KKT. poncu . KEY PRNEL DRAWING NO. TITLE Main Board Circuit Diagram, 8/9 MODEL NO. FX-2111 523-2020 DATE ASSEMBLY NO. PCB-6409 6-19 95.01.08 nxi 2 1 1 90 I . P2 n xg j i4qqi VDO 6 LCO-PCB Vtt VOO VCC L 1 2 3 0F3-1SS-2C 15 1 5 2 l-> 1 « 3 13 1 3 <sr 12 1 2 <R/M 1 1 1 1 10 10 g 9 <vS5 4 I - R /w _ 5 C ooo oe i 002 5 - 6 - 7 086 DB7 <e: 7 8 9 8 9 JO 1 I 1 1 12 1 2 13 1 3 1 <» 1 4 OB L 2 2 3 - <i b r ch i -rCLOCK 3 Page 1 6 POT POUT STROBES STW08CI THERMIST OR 8 DB 2 VDD 7 0B3 CNQ 6 0B4 CNO 5 PBS GNO 4 P85 0B7 connoN 3 3 COMnON 2 2 ConnpN 1 - LCD MODULE 1 5 s 6 6 7 7 B 8 9 9 T ST Ro6E - ' lO OPS CLOCK l " DB3 081 DOT ft r N PS 0F3«-15P-2DS« Cpgo-oBfl Page? S ISB-Ph-k.-S; THERMAL ph5 10 10 1 1 I 1 12 1 2 L3 L 3 14 1 4 IS 1 S tirnp - P .gp > 3 £ C467 0 1 T . E2 - <COMnON 1 P a<jr 7 Page 7 ~ iS - HEPD MODULE DRAWING NO. TITLE Main Board Circuit Diagram, 9/9 MODEL NO. FX-2111 523-2020 DATE ASSEMBLY NO. PCB-6409 6-21 95.01.08 J o o o u o n DD ? CD O KJ cn i».» c I? C i cnn Dm c I; > !;< Z r m ML/1 " n n 9 CP o D XI O 1 9 s2 o I s » o n B I (") C") » TITLE DRAWING NO. Power Board Circuit Diagram MODEL NO. FX-2111 1 B ASSEMBLY NO. 524-2021 DATE PCB-6410 6-23 96.01.08 r I s r> L 3 5 - a IN DRAWING NO. TITLE Sensor Board Circuit Diagram MODEL NO. FX-2111 524-1959 DATE ASSEMBLY NO. PCB-6239 6-24 94.07.14 NF-26P . /2'i . M 1 WED 2 r: i to n72 Or3i>-SP-2D50 n; 3 MT 4 _ ELU£ . s 6 MOTOR MODULE DRAWING NO. TITLE Motor Module Circuit Diagram MODEL NO. ASSEMBLY NO. 524-1960 DATE 94.07.14 FX-2111 6-25 VHfl-4M ©(- VHfi-ON H 1 2 9 6V . t HOOmOh 3 4 1 -I H H H Ni-MH I ' r BATTERY PRCK 6-HRflRFD DRAWING NO. TITLE Battery Pack Circuit Diagram MODEL NO. 524-1961 DATE ASSEMBLY NO. 94.07.14 FX-2111 6-26 F C P-2 2 2 2 A T INLET BOARD PI 0 PZ 0 0 P10 LCD Ni-MH i P2 BATTERY PACK i MOTOR P3 NF-26P1/25 DC [ N P4 PC8-6410 MOTOR CABLE MA I N THE RMAL HEAD THEMAL CABLE B 0 A RD i- \ P5 PATIENT P8 CABLE / Q56-F P6 P7 S W SENSOR BOARD PANEL F X - 2 1 1 1 PCB-6239 DRAWING NO. TITLE Overall Connection Diagram MODEL NO. ASSEMBLY NO. 433-5126 DATE 96.10.09 FX-2111 (USA) 6-27 FCP-2222W INLET BOARD PI i L CD : pio i Ni-MH BATTERY PACK i P2 u PCB-6410 v . u 6 P2 MOTOR P3 NF-26P1/25 DC TN P4 MOTOR CABLE MA IN THE RMAL THEMAL CABLE BOARD \ P5 s HEAD PATIENT P8 CABLE Q56-F P6 P7 SW SENSOR BOARD PANEL F X-2 1 1 1 PCB-6239 DRAWING NO. TITLE Overall Connection Diagram MODEL NO. ASSEMBLY NO. 4334894 DATE 96.01.08 FX-2111 (CE) 6-29 SOLDER NDM202AOO 14 3 2 > AXL214901 33 2 (7) m TO F4 TITLE DRAWING NO. LCD Connection Diagram MODEL NO. ASSEMBLY NO. 434-4895 DATE 96.01.08 FX-2111 6-31 MOTOR NF-26P1/2 5 r 10 SUMITUBE F2 3 xO. 25 SDMI TUBE F2 6x 0. 25 o HIRING (BRN) 1 0 0 mm HIRING SOLDER HIRING (RED) SOLDER UL1007 AHG28 6 HIRING LENGTH BRN connector RED ORG 1 7 0 contact YEL BLU PUR DF3-6S-2C 3 pin 4 pin Spin Spin DF3-24 28SC DRAWING NO. TITLE Motor Connection Diagram MODEL NO. 1 1 6 Omml pin 14 0 2 pin 434-4896 DATE ASSEMBLY NO. 96.01.08 FX-2111 6-32 CHAPTER 7 laqram Main Board Assembly Diagram, 1/2 & 2/2 433-4897 Inlet Board Assembly Diagram 434-4899 7-1 to 7-3 7-5 HOW 10 MOUNT IC22 (PQ30RV11) COMPONENT SIDE PCS-6X09 C35 SILK CHART is FLUX AREA . 65 rfi .:iiiiiiisiTnfria£ fp CI 6™ 5 ir ZD ? o su =s----flm?ji0 a z a n u-l 1 « 1 c D 9 IC4 EES) I lO"'" FUKUOA DEN5HI CO .. ( . . I * | , i-i how TOMOmr ic socket f en PCB-6409 lr_ LTD. i - "li-=nt-rJ rn0 31 w 5 w - n 0.5 (ZDc _ 2 ICIO 225 AT D CI HOW TO MOUNT 04 .at .2 2 (2SJ245) M n B1 1 3 . 2J B2< = x x X 1 I I 1 - - s [v' Pi LED ys' L O m AREA : 5 PC BOARD OPEN MAX (MOUNT icia 3 I .i-l IC1 :512K IC1 : 1M) Li - HOW 7t?MQUNT Cft? CECHEW105JZ) LEAD CUT LIMIT MAX 1 5 mm . St IC31 DRAWING NO. TITLE Main Board Assembly Diagram, 1/2 PC BOARD MODEL NO. FX-2111 ASSEMBLY NO. 433-4897 DATE PCB-6409 OMPOfCNT SIDE 7-1 95.12.15 is FLUX AREA . 1 5 tV jr'' W AREA ARE NOT REQUI m TITLE DRAWING NO. Main Board Assembly Diagram, 2/2 MODEL NO. FX-2111 DATE ASSEMBLY NO. PCB-6409 0 m > X) X > CD x o 6 CO o 33 5 CO > m ft s t CD 5 CD 4 O LTI CD O < ro 30 - \ m o 6 2 O O o o o 3 O cz -i mu be z 3 -I CO m TITLE DRAWING NO. Inlet Board Assembly Diagram MODEL NO. FX-2111 434-4899 DATE ASSEMBLY NO. PCB-6410 7-5 96.01.08 CHAPTER 8 Electrical Parts Lists 1 FX-2111 Main Blocks 674-11338 8-1 2 Main Board, PCB-6409 SAS SMD for FX-2111 674-11339 8-2 3 Main Board, PCB-6409 THD for FX-2111 674-11340 8-9 AC Inlet Board, PCB-6410 THD for FX-2111 (115V)... 674-11664 8-11 AC Inlet Board, PCB-6410 THD for FX-2111 (230V)... 674-11340 8-12 LCD Assembly for FX-2111 Parts Supplied to Wiring/Assembling Process 674-11342 8-13 674-11343 8-14 . . . 4 . 5 . 6 . 7 . for FX-2111 1 FX-2111 Main Blocks . 674-11338 Parts No. 9E3041 Description Symbol Drawing No. PC Board AS Q'ty 1 Main Board 1 9F3419 PCB-6409 with ROM 9F3293 ROM Set 9F3424 PCB-6409 SAS 9G3739 PCB-6409 SMD 674-11339 1 9G3740 PCB-6409 THD 674-11340 1 674-10486 1 1 AC Inlet Board 9F3420 PCB-6410 Assembly 674-11341 1 9F3421 LCD Assembly 674-11342 1 9D0995 Set of Supplied Parts to Assembling Process (CE) 9E2793 Set of Assembled Electrical Parts (2307) 674-10513 1 9E3042 Set of Assembled Electrical Parts (CE) 674-11359 1 9D0996 Set of Supplied Parts to Wiring Process (CE) 674-11343 1 8-1 1 Remarks 2 Main Board, PCB-6409 SAS SMD for FX-2111 . 674-11339 Parts No. Drawing No. Description Symbol Q'ty Resistors 1A7510A R1, 15, 23, 75-77, 145, 163, 168, 176, 181, 192, 197, 200, 202, 207, 209, 212, 219, 220, 226, 227, 229, 230, 232, 236, 239, 242, RK73H2AF, Square Chip lOkfi 30 RK73H2AF, Square Chip 20kfi 51 245 1A7503A R2~11, 14, 26, 29, 30, 34-40, 43, 46 57, 68-70, 86-92, 140, 158, 182, 183, 189-191, , 195, 201, 210, 225, 196, 204, 211, 241, 254, 255 1A7685A R12, 84, 110 RK73H2AF, Square Chip 12kG 3 1A9662A R13, 41, 51, 72, 80, 199, 206, RK73H2AF, Square Chip 30kn 8 RK73H2AF, Square Chip 560kn 3 RK73H2AF, Square Chip 200kn 8 RK73H2AF, Square Chip 3kil 5 RK73H2AF, Square Chip 1kfi 14 215 1A7641A R16, 117, 148 1A9666A R17, 32, 58, 116, 169, 173, 231, 252 1A9657A R19, 59, 119, 164, 222 1A9655A R20, 28, 64, 65, 67, 82, 93-100, 105, 109, 132, 133, 136, 144, 150, 156, 217, 248 8-2 Remarks 2 Main Board, PCB-6409 SAS SMD for FX-2111. continued . 674-11339 Parts No. Symbol Description Drawing No. Q'ty 1A7512A R85, 147 RK73H2AF, Square Chip 16kn 2 1A7502A R22 RK73H2AF, Square Chip 6 8kG 1 1A9263A R24 SR73K2EJ, Metal Film 0 1fi/0.5W 1 1A7522A R25, 63, 81, RK73H2AF, Square Chip lOOkfi 5 RK73H2AF, Square Chip 5 1k0 10 . . 141,142 1A7545A R27, 61, 103, 171, 174, 214, 216, 228, . 249, 250 1A7571A R31 RK73H2AF, Square Chip 300fl 1 1A9631A R33 RK73K2AF, Square Chip 100 1 1A7534A R42, 62 RK73H2AF, Square Chip 3 48kU 2 1A7584A R44, 45, 56, 78, 124, 125, 165, RK73H2AF, Square Chip 1Mfi 11 RK73H2AF, Square Chip 1001] 6 RK73H2AF, Square Chip 2kU 14 166 , . 218, 244, 246 1A9653A R47, 48, 55, 74, 115, 161 1A9656A R49, 50, 52, 71, 73, 126-129, 143, 159, 178, 179, 187 1A9396 R53 RN73F2ATD-B, Metal Film 1 1A9398 R54 RN73F24TD-B, Metal Film 1 1A9303A R60 RH73H2AF, Square Chip 28.7k(] 1 1A9271 R66 RK73H3AF, Metal Film IkO/IW 1 1A7681A R79 RK73H2AF, Square Chip 2 2kQ 1 R7519A R83, 146, 160, 193, RK73H2AF, Square Chip 51 kl) 15 198, 205, 213, 223, 240, 203, 208, 221, 237, 243, 247 8-3 . Remarks 2 Main Board, PCB6409 SAS SMD for FX-2111., continued . 674-11339 Parts No. Symbol Drawing No. Description Q'ty 1A7547A R85, 147 RK73H2AF, Square Chip 16kn 2 1A7656A R101, 151, RK73H2AF, Square Chip 240ki] 3 162 1A7558A R104, 122 RK73K2AF, Square Chip 42.2fl 2 1A7689A R106, 108, 111, 155, RK73H2AF, Square Chip 82kn 5 167 1A9665A R107 RK73H2AF, Square Chip 120kQ 1 1A9397 R112~114 RN73F2ATD-B, Metal Film 78.7ki] 3 1A7674A R118, 121, 123, 149, RK73H2AF, Square Chip 430n 5 RK73H2AF, Square Chip 13kn 4 152 1A7652A R120, 131, 157, 238 1A7616A R130 RK73H2AF, Square Chip 56kn 1 1A9270A R134 SR73H2EF, Metal Film 0 51Q/0.25W 1 1A7604A R135 RK73H2AF, Square Chip 1 43k£l 1 1A7540A R137, 138, RK73H2AF, Square Chip 24kQ 4 . . 233, 251 1A7554A R139 RK73H2AF, Square Chip 36kfi 1 1A7578A R153, 154 RK73H2AF, Square Chip 150fi 2 1A7202 R170 TRN60LG, Cylindrical Tantalum Alloy 1A7591A R175, 180, RK73H2AF, Square Chip 3 5Mn 1 160kfi 4 . 234, 253 1A7526A R172, 177 RK73H2AF, Square Chip 1A9393 R184 RN73F2ATD-B, Metal Film 11011 1 1A7640A R185, 186 RK73H2AF, Square Chip 140kfi 2 1A7587A R188 RK73H2AF, Square Chip 3 4kQ 1 1A9395 R194 RN73F2ATD-B, Metal Film 13kn 1 1A7668A R256~259 RK73H2AF, Square Chip 220n 4 1A7528A R224 RK73H2AF, Square Chip 698kH 1 2 8-4 . Remarks 2 Main Board, PCB-6409 SAS SMD for FX-2111, continued . 674-11339 Parts No. Description Symbol Drawing No. Q'ty Capacitors 1B5021A 01,2, 4, 6, 7 11, 13, 14, 22, 25, 26, 28-34, 36, 37, 39, 40, 42, 43, GRM40F104Z50, Layered Ceramic Chip 0.1 pF 59 , 45, 49, 50, 52, 53, 61 -63, 66-68, 73, 76, 77, 85, 86, 89, 93, 94, 96, 97, 99, 100, 105-108, 116-119, 125-128 1B1701 C3 ECSTCY105, Tantalum Electrolytic 1.0(jF 1 1B5412A C5, 27 EEV-H81CV220, Aluminum Electrolytic 22nF/16V 2 1B5408A C8 EEV-HBIHV0010, Aluminum Electrolytic 1mF/50V 1 1B5073A C9, 16, 41, 56, 58, 87, GRM40CH1U2J50, Layered Ceramic 7 lOOOpF 88 1B0702 1B5055A C10 C12, 51, 69, 25CV100GX, Aluminum Electrolytic Chip 100uF/25V for High Frequency 1 GRM40B222K, Layered Ceramic 2200pF 4 134 1B0209 C15 16SM6R8M, OS 6.8mF/16V 1 1B0208 C17, 21 10SM22M, OS 22tiF/10V 2 1B5019A C18 GRM40CH101J, Layered Ceramic 100pF 1 1B5022A C18, 20, 57, 59, 60, 65, GRM40B103K, Layered Ceramic O.OIpF 8 MCH215C223K, Layered Ceramic 0.022mF 5 51pF 1 75, 115 1B5062 C23, 24, 38, 47, 71 1B5002A C35 GRM40CH510J, Layered Ceramic 1B8501 C44, 90, 91 35CV47GX, Aluminum Electrolytic Chip for High Frequency 47pF/35V 1B5815 C46 ECHU1H102G, Layered Film 8-5 1000pF 3 1 Remarks 2 Main Board, PCB-6409 SAS SMD for FX-2111 . 674-11339 Parts No. 1B5027 Description Symbol C48, 131, Drawing No. MCH215A561J, Layered Ceramic 3 3300pF Q'ty 133 1B5816 C54 ECHU1C332G, Layered Film 1B5402A C64, 70, 95 EEV-HB1 EV4R, Aluminum Electrolytic 4 7nF/25V 3 . 1 1B5513A C55, 83 GRM42-6F224Z, Layered Ceramic 0.22nF 2 1B5011A C72, 74 GRM42-6B473K, Layered Ceramic 0.047mF 2 1B5818 C78 ECHU1C683J, Layered Film 0.068mF 1 1B5261 C79, 80, 92 ECWU1C224J, Layered Film 0.22nF 3 1B5076A C81, 100, 110, 112, GRM40B332K, Layered Ceramic 3300pF 6 124, 135 1B5044A C84 GRM40B153K, Layered Ceramic 0.015nF 1 1B5003A C98, 103, 104, 113, 114, 120, 121, 123, GRM40CH101J, Layered Ceramic 150pF 9 129 1B5403A C101, 102 EEV-HB1EV330, Aluminum Electrolytic 4 7nF/16V 2 . 1B5808 C111,113 ECHU1H140J, Layered Film O.lpF 2 1B5014C C130 MCH325F274Z, Layered Ceramic 0.47nF 1 1B5001A C132 GRM40CH300J, Layered Ceramic 30pF 1 1B5074A C138 GRM40CH331J, Layered Ceramic 330pF 1 Diodes 3929 D1~6 D3FS4A (2.6A, 40V) 6 1D4200 D7 02CZ-5.1V, Zener 1 1B3885 D8, 1, 15, 32 1SS301 3 1D3886 D9, 11, 12, 16, 17, 19, 20, 23-31, 1SS302 18 33, 34 1D4113 D10 02CZ-3.0K, Zener 1 1D3928 D13, 14, 18, D1FS4A 4 21 8-6 Remarks 2 Main Board, PCB-6409 SAS SMD for FX-2111, continued . 674-11339 Parts No. 1D4552 Description Symbol D22 U1ZB12, Zener Drawing No. Q'ty 1 Transistors 1D1735 Q1~3, 5, 8, 12 14, 15, 27, 29, 30, UN5212 13 , 33, 38 1D989 Q6 2SC-2713GR 1 1D0116 Q7 2SA-1163G 1 1D2385 Q9 2SK-739-2, FET 6 1D2392 Q10 2SK-1062, FET 1 1D0117 Q11 2SA-1162V 1 1D1734 Q13, 18, 26, 28, 42, 43 UN5112, Digital 6 1D0968 Q16 2SC-1623(L5/L6) 1 1D2394 Q17 2SJ245S, FET 1 1D2382 Q19, 22, 23, 31, 32, 34 SST4393, FET 6 1D2342 Q20, 21 2SK1920-FA, FET 2 1D0153 Q25 2SA-1463(1 K) 1 1D1713 Q35~37, 39-41, FMW2 9 44-46 Integrated Circuits 1E5037A IC2, 5 HM6264BLFP-10LT, RAM 2 1E2769 IC4 NJM2103M, Reset IC 1 1E6168 IC6 HA16120FP, Switching Regulator 1 1E6245 IC7 PQ05SZ5, Low-loss Regulator 1 IE2621 ICS FD88007-AC, Gate Array 1 1E6235 IC9, 33 NJM78L05UA, 3-terminal Regulator 2 1E7205 IC10 S-8437AF 1E5390A IC11 TC-74HC245AF, CMOS Logic , Inversion-type Switching Regulator 8-7 1 1 Remarks 2 Main Board, PCB-6409 SAS SMD for FX-2111, continued . 674-11339 Parts No. Drawing No. Description Symbol Q'ty 1E4424 IC12, 19 TC-4584BF, CMOS Logic 2 1E0374 IC13 HPC4074G2, Low-noise FET-input OP Amp. 1 1E5397A IC14, 23 TC-74HC04AF, CMOS Logic 2 1E5412A IC15 TC-74HC541AF, CMOS Logic 1 1E6167 IC18 bq2003S> Quick Charging IC 1 1E5543 IC16 TC7W04F, L-MOS Logic 1 1E4169 IC17 TC-4053BF, CMOS Logic 1 1E5595 IC20 TC-74AC74F, CMOS Logic 1 1E0198 IC21, 31 pPC4064G2, Low Power CoSsumption J-FET Input Operational Amplifier 1E0715 IC24 pPC4064G2(2), Low Power Oonsumption J-FET Input Operational Amplifier 1E2613A IC25 HEF4046BT, PLL 1 1E0266 IC26 MPC358G2, Comparator 1 1E5542 IC27 TC7W02F, L-MOS Logic 1 1E8961 IC28~30 PS2652L2-V(K), Photo Coupler 3 1E2083 IC32 LT1004CS8-1.2, Voltage Reference 1 1E7221 IC34 NJM79L05UA, 3-terminal Regulator 1 1E4177 IC35 TC-4093BF, CMOS Logic 1 1E4170 IC36, 38, 41, TC-4051BF, CMOS Logic 4 45 1E0264 IC37, 39, 42 MPC4064G2(2), Low Power Oonsumption J-FET Input Operational Amplifier 1E4423 IC40 TC-4535BF, CMOS Logic 1 1E0365 IC43 pPC339G2, Comparator 1 1E4182 IC44 TC-4094BF, CMOS Logic 1 Coil 1T6047 1H6409 L2 470mH LPC4045TE-471K Printed Circuit Board, PCG-6409 8-8 1 1 Remarks 3 Main Board, PCB-6409 THD for FX-2111 . 674-11340 Description Symbol Parts No. Drawing No. Q'ty Resistor 1A7040 R18 RSSX3L20 0.01J/3W, Metal Film 1 Capacitors 1B4701 C82 ECHE1H105JZ, Metalized 1 1B0866 IC136 URS1E222MHZ 2200/25V, Aluminum Electrolytic 1 1B0713 CI37 URS1V102MHZ 1000/35V, Aluminum Electrolytic 1 Transistors 1D2393 Q4, 24 2 2SJ245L, FET Crystal Oscillator 1D7620 XI 1 JX0-5S 16MHz (with stand) Integrated Circuits 1E1957 IC3 HD63B09E, CPU 1 1E7021 IC22 PQ30RV1/11, Low-loss Regulator 1 1E9198 DIC«?iG11S1 IC Socket 1 1E9199 DIC640G11S1, IC Socket 1 , Lithium Battery 1U0231 BT1 CR-1/3-FT2-1 684-2428 1 Connectors 1F0215 P1 B3P-VH 1 1F1068 P2 AXB114001 BB 1 1F3208 P3 S-G8036 #01 1 1F0562 P4 DF3A-6P-2DSA 1 1F9376 P5 DF3A-15P-2DSA 1 1F0548 P6 008370061000800 1 1F1131 P7 008370151000800 1 1F1004 P8 RDAG-15SE1(F) 1 8-9 Remarks 3 Main Board, PCB-6409 THD, continued . 674-11340 Parts No. 1F0365 P10 Drawing No. Description Symbol B4P-VH Q'ty 1 Arrestor 1L4807A NE1 Y06S-230B 1 Fuses 1L4057 F1, 2 3 15A TR-5 K19374 2 . Coils & Ferrite Core 1T6242 L1 HK-05S035-2010 RBP 200mH 684-2109 1 1T6243 L3 HK-04D030-1510 RBP 150nH 684-2108 1 1L4584 L4 BL01RN1-A62, Ferrite Core 1 Isolation Transformers 1T6015A T1 FT-2111 684-2153 1 1T6016A T2 FT-2155 684-2154 1 Piezoelectric Buzzer 1S3200 BZ1 1 PKM13EPP-4002 8-10 Remarks 4 . AC Inlet Board, PCB-6410 THD for FX-2111 (115V) 674-11664 Parts No. Symbol Description Drawing No. Q'ty Fuses 1L4476 239001 1L4905 MF-561A, Fuse Holder 1A 2 2 Locker Switch 1G3508 SJ-2S4A-07BB 1 AC Inlet 1D2393 NC176-1.5 1 GND Terminal 1K1151 FX-2111GND (A) 1 Connector 1F0159 1 B2P-VH Choke Coil 1T6342 TF1028S-801Y2R0-01 1 Spacer 1Z4311 1/2 LR-1.5 Printed Circuit Board 1H6410 323-3961 PCB-6410 8-11 1 Remarks 5 . AC Inlet Board, PCB-6410 THD for FX-2111 (230V) 674-11340 Parts No. Drawing No. Description Symbol Q'ty Fuses 1L4091 No. 19195 1L4905 MF-561A, Fuse Holder 500mA 2 2 Locker Switch 1G3508 1 SJ-2S4A-07BB AC Inlet 1D0288 NC176-1.5 1 GND Terminal 1K1151A 314-3471 GND (A) 1 Connector 1F0159 1 B2P-VH Choke Coil 1T6342 1 TF10283-801Y2R0-O1 Spacer 1Z4311 1/2 LR-1.5 Printed Circuit Board 1H6410 1 PCB-6410 8-12 Remarks 5 . LCD Assembly for FX-2111 674-11342 Parts No. Symbol Description Drawing No. Q'ty LCD 1 LI 709 NDM202A00 1 Connector 1F1085 AXL214901 1 Spacers 6B8308 SP-15 2 6B9216 Grounding Piece 1 Shielding Fingers 6B9217 (A) 1 6B9218 (B) 1 8-13 Remarks 7 . Parts Supplied to Wiring/Assembling Process for FX-2111 674-11343 Parts No. Description Symbol Drawing No. Q'ty 434-4725 1 1M0385 Motor Assembly 1M0466 Inlet Board Grouding Cable 3 with Core 1M0482 Inlet Board Grounding Cable 3 1L4581 Ferrite Core, 5977-000601 1 6G0234 Heat Shrink Tube, F2 25x0.4, 30mm long 1 1K0031 Press-fit Terminal, TMEV1.25-3 1 8-14 1 684-2415 1 Remarks CHAPTER 9 Spare Parts Lists PCB-6238 2 PCB-6236 674-10554 674-10555 9-1 9-2 Paper Magazine 674-10556 9-3 Recorder 674-10557 9-4 Upper Case 674-10558 9-5 Lower Case 7 Screws 674-10559 674-10560 9-6 9-7 1 . . 3 . 4 . 5 . 6 . . 1 PCB-6238 . 674-10554 Parts No. Symbol Description Drawing No. Q'ty 9E2963 PCB-6238B (2) Assembly for 115V 9F3331 PCB-6238 Assembly for 115V 5T3111 Nuts, Type 3 M6 2 5T2009 Spring Washer, Nominal Dia. 6mm 1 9E2964 PCB-6238B (3) Assembly for 230V 1 9F3310 PCB-6238 Assembly for 230V 1 5T3111 Nut, Type 3 M6 2 5T2009 Spring Washer, Nominal Dia. 6mm 1 1L4476 Fuse, 239001 for 115V 2 1L4475 Fuse, 239500 for 230V 2 9-1 1 Remarks 2 PCB-6236 . 674-10555 Parts No. Symbol Description Drawing No. Q'ty 9E2965 PCB-6236 (1) Assembly 1 9F3307 PCB-6236 Assembly for FX-2111(N) 1 5R1604 Double Semus Screw, M3x6 2 9E2966 PCB-6236 (2) Assembly 1 9F3309 PCB-6236 (with ROM) for FX-2111 (N) 1 5R1604 Double Semus Screw, M3x6 2 1E8531 Programmed ROM for FX-2111 (36-202) 1 1L4057 Fuse, TR-5 K19374, 3.15A 2 1U0231 Lithium Battery, CR-1/3-FT2-1 684-2428 1 9F3305 LCD Assembly 634-4726 1 9F3305 Spacer, SP-15 2 9-2 Remarks 3 . Paper Magazine 674-10556 Parts No. Symbol Description Drawing No. Q'ty 9H0402 Platen Roll Assembly 6B8324A Platen Roll 6A5350 Bearing, LF-740ZZ 6B8325 Gear, 38A 5R3003 Countersunk Screw, M2x6 1 9F2648 Paper Magazine Assembly 1 6B8323 Paper Magazine 9H0402 Platen Roll Assembly 1 224-0890 1 2 224-0313 222-0889 1 1 1 9-3 Remarks 4 Recorder . 674-10557 Parts No. Description Symbol Drawing No. Q'ty 9H0617 Gear 38B Assembly 6B8320 Gear, 38B 5R8001 Setscrew, M3x3 1 9H0812 Thermal Array Head Assembly 1 6B8316 Thermal Array Head, Q56-F 684-2092 1 6B8316A Thermal Array Head Mounting Plate 213-3448 1 6B8394 Tension Coil Spring, DE542 2 5R0602 Semus Screw, M3x5 2 9F3354 Recorder Assembly 1 9H0812 Thermal Array Head Assembly 1 9H3215 Recorder Chassis SU 1 1M0385 Motor Assembly 1 9H0617 Gear 38B Assembly 1 9H3216 Grounding Piece (C) SU 6B8307 Grounding Flexible Conductor 6B8395 Sensor Presser Sponge 9F3289B PCB-6295SMD Sensor Board 1M0381 Thermal Array Head Cable 5R1851 Double Semus Screw, M3x4 1 5R1604 Double Semus Screw, M3x8 1 9H3522 Paper Shaft SU 1 244-0314 313- 3461 1 6B8307 1 314- 3465 1 674-10490 1 684-2093 1 224-0891 9-4 1 1 Remarks 5 . Upper Case 674-10558 Parts No. Description Symbol Drawing No. Q'ty 1 9F3355 Upper Case Assembly 9H3211C Upper Case SU 111-6621 1 6B8305 LCD Filter 114-6625 1 6B9171 Lock Spring 114-6811 1 1G9340A Key Panel 684-2146 1 5H9128 Rated Power Label (B) for 115V 154-4573 1 5H9130 Rated Power Label (C) for 230V 154-4575 1 9-5 Remarks 6 Lower Case . 674-10559 Parts No. Symbol Description Drawing No. Q'ty 9F3356 Lower Case Assembly (1) 9H3212C Lower Case SU 111-6622 1 6B8306 Insulation Sheet 314-3463 1 6B8327 Batter Presser Sponge 314-3466 1 9H3213A Grounding Piece (A) SU 313-3459 1 1T1558 Transformer, FCP2222AT for 115V 684-2158 1 1T1559 Transformer, FCP-2222UT for 230V 684-2157 1 6B8326A Battery Cover 112-6624 1 5H9127 Instructions Label (B) 154-4572 1 5H9138 Instructions Label (C) for 230V 154-4577 1 5H9129 Fuse Label (B) for 115V 154-4574 1 5H9131 Fuse Label (C) for 230V 154-4576 1 5H5639 Lithium Battery Label 154-3574 1 6B6796 TM-166 No. 12 1 1 9-6 Remarks 7 Screws . 674-10560 Parts No. Symbol Description Drawing No. Q'ty 9H3238 Set of Screws for Ni-MH 5R1604 Double Semus Screw, M3x8 5R0602 Semus Screw, M3x5 2 5R1851 Double Semus Screw, M3x4 1 5R3003 Countersunk Screw, M2x6 1 5R8001 Setscrew, M3x3 1 9-7 1 24 Remarks CHAPTER 10 Structural Diagrams FX-2111 (USA) External Appearance (1) External Appearance (2) 413-0994 413-0995 10-1 10-3 424-3619 10-5 423-3620 423-3621 422-3622 422-3623 412-0996 10-7 10-9 10-11 10-13 10-15 413-0849 413-0927 10-17 10-19 AC Inlet Board 424-3443 10-21 Main Board* 423-3444 10-23 Paper Magazine Assembly* Paper Shaft Assembly* Recorder Assembly* Upper Case Assembly Lower Case Assembly (1) Lower Case Assembly (2) Lower Case Assembly (3) Exploded View 423-3250 424-3251 423-3252 423-3445 423-3446 423-3447 422-3448 412-0928 10-25 10-27 10-29 10-31 10-33 10-35 10-37 10-39 AC Inlet Board Upper Case Lower Case Assembly (1) Lower Case Assembly (2) Lower Case Assembly (3) Exploded View FX-2111 (CE) External Appearance (1) External Appearance (2) > Asterisked diagrams are common to both FX-2111 (USA) and FX-2111 (CE). 2 6 0 3sg II 8 .S5 ii A Of as TO 5 [J a DRAWING NO. TITLE External Appearance (1) MODEL NO. ASSEMBLY NO. 413-0994 DATE 96.09.30 FX-2111 (USA) 10-1 JUL JUL © 0) Q ClITIOI-Ti I«I.t r ii»n iti(tci«| til It r i 11 if dtclric ibid, ti t*tIi(i «d if f i 111 ul rii*Tc tirtr. firiooi. 1, D DASOBI-liil *f ciflniiB flmtblt <1 DA NG B K - 1 i ii it 9 o - if ttti Ii ' pruiii* d lBtltkttilstt D 51 10DE 1ISBT V <3 0 taV @ ® ei 2jnv Mil IS @ A o a STOP 3J 7 dpUr*' ilflilBikltl. i HAD START DFF the grtietct tf ruse $1 C A R D I MA X Dl pn HIJITIIltT [ Vrjcuwdensh) Pi- 2 111 ii ittilkitici. d'tiplntoi. 2 6 0 E CA8T10R:FBD8RAl Ul SISTIICTS T8IS DtVICt TO SALS BT 01 01 71! 0IDI1 01 A PB1S1CIA1. - 1 Tizr 6 5mm DRAWING NO. TITLE External Appearance (2) MODEL NO. ASSEMBLY NO. 413-0995 DATE 96.09.30 FX-2111 (USA) 10-3 2 \ a* o ® o 0 O 2-M8 Nut & Plain Washer T r Key No. Parts No. 1 1L4476 2 1K1151A i Description Drawing Na 314-3471 Fuse, 239001 2 GND Terminal (A) 1 AC inlet Board FX-2111 (USA) Remarks DRAWING NO. TITLE MODEL NO. Q'ty 424-3619 DATE ASSEMBLY NO. PC 6-6410 10-5 96.09.30 5 3 i 6 '' 1 A - 55 33 0 2 ti 1 »l 4 Si'S II s Of S»!|!Li TO 8 7 Key No. Parts Na Description Drawing Na 1 Q'ty Upper Case Assembly 1 2 9H3494 111-6862 Upper Case Assembly SU for USA 1 3 1G9340A 684-2146 Key Panel 1 4 6B8305 114-6625 LCD Filter 1 5 5H9128 154-4573 Rated Power Label (B) 1 6 6B9171 114-6811 Lock Spring 1 7 5H9237 154-4670 Caution Label 1 Remarks DRAWING NO. TITLE Upper Case MODEL NO. ASSEMBLY NO. 423-3620 DATE 96.09.30 FX-2111 (USA) 10-7 4 2-Double Semus Screw M3x8 , 2-Taping Pan-head Screw M2.6x6 , 6 A A A C 3 Pasted 2 A C f_ o TV 1 Key Na Parts Na Description Drawing Na 1 Q'ty Lower Case Assembly (1) 1 2 9H3495 111-6863 Lower Case Assembly SU for USA 1 3 6B8327 314-3466 Battery Presser Sponge 1 4 6B8326 112-6624 Battery Cover 1 5 5H9374 154-4669 Fuse Label (D) 1 6 9E7329 Rating label 1 Remarks DRAWING NO. TITLE Lower Case Assembly (1) MODEL NO. ASSEMBLY NO. 423-3621 DATE 96.09.30 FX-2111 (USA) 10-9 7 Key No. Drawing No. 1 B A Parts No. 3 Q'ty Lower Case Assembly (1) 1 423- 3621 Lower Case Assembly (1) 1 9H3213A 313- 3459 Grounding Piece 1 684-2156 Transformer, FCP2222AT 1 2 4 Description 4 1T1558 5 9F3420 424- 3619 PCB-6410 Assembly 1 6 6B8306 314- 3463 Insulation Sheet 1 7 6B9198A 314-3601 Insulation Cover 1 Remarks \ .v A 3 B O 5 2 o 3 o 9 a 0 6 o o DRAWING NO. TITLE Lower Case Assembly (2) MODEL NO. ASSEMBLY NO. 422-3622 DATE 96.09.30 FX-2111 (USA) 10-11 2 9-Double Semus Screw M3x8 , G c 4 E F IF D 0 . c D A G it E 5 B O 3 F O 5 0 11 0 -1 p 6 V o Key Na Parts Na 1 Description Drawing Na Q'ty 422- 3622 Lower Case Assembly (2) 1 2 9F3354 423- 3252 Recorder Assembly 1 3 1M0381 684-2093 Thermal Array Head Cable 1 4 1T1558 423-3444 PCB-6409 Assembly 1 Remarks 7 DRAWING NO. TITLE 5 5H9615 154-4836 Lead Label (C) 1 6 5H5639 154-3574 Lithium Battery Label 1 7 5H9417 154-4821 Patient Label 1 Lower Case Assembly (3) MODEL NO. ASSEMBLY NO. 422-3623 DATE 96.09.30 FX-2111 (USA) 10-13 2 i 5S> ' II D At r A m 3* Li 6 A Double Semus Screw 3 M3x8 4 A B ii s" 2-D-jubl e Semus Screw M3x8 11 , Key Na Parts No. Description Drawing Na Q'ty Remarks DRAWING NO. TITLE 1 423-3623 Lower Case Assembly (3) 1 2 423-3620 Upper Case Assembly 1 423-3250 Paper Magazine Assembly 1 3 4 9F2648 9H3522 224-0891 Paper Shaft SU 1 Exploded View MODEL NO. ASSEMBLY NO. 412-0996 DATE 96.09.30 FX-2111 (USA) 10-15 o 2?5 v . >5- \ DRAWING NO. TITLE External Appearance (1) MODEL NO. ASSEMBLY NO. 413-0849 DATE 94.10.07 FX-2111 (CE) 10-17 - I [ I @ /Ft, o Cilil01-Ti Itfir TtAtEi itttitin tki 11 nil . »f ilttltit tiflid tkit nrvlci t* tit rtiif* ititt. pt r 11 d i * 1. u DkHQtl'litl If tlfltiitt f 11 mm t k It if tie* ii tk* frimci sf 0 (inlbitiei. D A B C E S -1 i n»« *'h|Iiii»i. It in tMplertf > ptititu d inttl«*Miii iifliaiiMn. o o n S I 2»TI 5i oo 1DDI UISITITITT 5 I ; CA R D I MA X FX-Z11 J IBID ' 11 oil:?: " STAKT PIT OTf V A STOP Y 3J v 0 0 !=. ii ] - si I o a . 2 6 0 nzr DRAWING NO. TITLE External Appearance (2) MODEL NO. ASSEMBLY NO. 413-0927 DATE 96.01.24 FX-2111 (CE) 10-19 2 \ 0 ® 0 O o 2_M8 Nut & Plain Washer T T Key No. Parts Na 1 1L4475 2 1K1151A 1 Drawing Na 314-3471 Description Fuse, 239500 2 GND Terminal (A) 1 TITLE FX-2111 (CE) Remarks DRAWING NO. AC Inlet Board MODEL NO. Q'ty 424-3443 DATE ASSEMBLY NO. PCB-6410 10-21 96.01.24 5 2-Binding Screw M3x8 , 7 r CD 2 .j 3 o o 3 r 0 v. \ 6 y y y 0 y y y 4 y y y y y 0 y 2-Double Semus Screw M3x8 , y y Key No. Parts No. Drawing No. Description Q'ty 1 9F3424 PCB-6409 SAS 1 2 9F3419 PCB-6409 with ROM 1 3 1L4057 Fuse, TR-5 K1937, 3.15A 2 4 1U0231 Lithium Battery, CR-1/3N-FT-2-1 1 5 1E8631 Programmed ROM, FX-2111 (36-202) 1 6 9F3421 LCD Assembly 1 7 1L1709 LCD, NDM202A0O 1 8 6B8308 Spaeer, SP-15 2 9 6B9216 314-3602 Grounding Piece (D) 1 10 6B9217 314-3603 Shielding Finger (A) 1 684-2428 Remarks TITLE 11 6B9218 314-3604 Shielding Finger (B) 1 DRAWING NO. Main Board MODEL NO. ASSEMBLY NO. FX-2111 (CE) 423-3444 DATE 96.01.24 10-23 5 Countersunk Screw M2x6 , 1 - J6 - - J 1 4 2 3 Key Parts No. Drawing No. 1 6B8323 222-0889 2 9H0402 Na 3 6B8324A 4 6A5350 5 6B8325 Platen Roll 224-0313 Description Paper Magazine 1 Platen Roll Assembly 1 Remarks 1 LF-740ZZ Bearing 2 Gear, 38A 1 DRAWING NO. TITLE Paper Magazine Assembly MODEL NO. Q'ty ASSEMBLY NO. 423-3250 DATE 94.10.01 FX-2111 10-25 2 1 Key Description Q'ty Parts Na Drawing Na 1 6B8321 223-0892 Paper Shaft 1 2 6B8322 224-0893 Back Tension Spring 1 No. DRAWING NO. TITLE Paper Shaft Assembly MODEL NO. Remarks ASSEMBLY NO. 424-3251 DATE 94.10.07 FX-2111 10-27 2-Double Semus Screw M3x5 , 5 3 6 D 1 1 7 4 Setscrew, M3x3 A A ) a «1 v . 0 j c Key Na 1 1 9 Double Semus Screw M3x4 , C 2 D 3-Double Semus Screw, M3x8 Parts Na Drawing Na Description Q'ty 1 9H3215 211-3447 Recorder Chassis SU 1 2 1M0385 434-4725 Motor Assembly 1 3 9H0617 Gear 38B Assembly 1 4 6B8320 Gear 38B 1 5 9H0812 Thermal Array Head Assembly 1 6 6B8316A Thermal Array Head Mounting Plate 1 7 6B8317 Tension Coil Spring, DE511 1 224-0314 213-3448 Thermal Array Head Q56-F 1 PCB-6239 SMD, Sensor Board 1 313- 3461 Grounding Piece C SU 1 6B8395 314- 3465 Sensor Presser Sponge 1 6B8307 314-3464 Flexible Grounding Conductor 1 8 1W3006 684-2092 9 9F3289B 674-10490 10 9H3216 11 12 , TITLE Remarks DRAWING NO. Recorder Assembly 423-3252 Double Semus Screw, M3x8 MODEL NO. ASSEMBLY NO. FX-2111 DATE 94.10.07 10-29 7 3 6 II II It A .St .4> 0 2 4 Key Na Parts Na Drawing Na 1 9F3355 2 9H3211C 111-6621 Description Q'ty Upper Case Assembly 1 Upper Case Assembly SU 1 3 1G9340A 684-2146 Key Panel 1 4 6B8305 114-6625 LCD Filter 1 5 5H9130 154-4575 Rated Power Label (C) 1 6 6B9171 114-6811 Lock Spring 1 7 5H9428 154-4694 CE Marking Label 1 Remarks TITLE DRAWING NO. Upper Case Assembly MODEL NO. ASSEMBLY NO. FX-2111 (CE) 423-3445 DATE 96.01.24 10-31 b 2-Double Semus Screw , M3x8 A 5 rJ5 A A 6 c 4 <S> Pasted ® . 2 r A C <s> 1/ OS v Key No. Parts Na Description Drawing Na ' Q ty 1 9F3356 Lower Case Assembly (1) 1 2 9H3212D 111-6622 Lower Case Assembly SU 1 4 6B8327 314-3466 Battery Presser Sponge 1 5 6B8326 112-6624 Battery Cover 1 6 5H9131 154-4576 Fuse Labell 7 9E7293 Remarks TITLE Rating Label Lower Case Assembly (1) MODEL NO. 1 DRAWING NO. ASSEMBLY NO. FX-2111 (CE) 423-3446 DATE 96.01.24 10-33 7 Key Parts Na No. 1 Drawing Na 9F3356 3 9H3213A 313- 3459 4 1T1559 684-2157 Transformer FCP2222UT 5 9F3420 424- 3443 PCB-6410 Assembly 1 6 6B8306 314- 3463 Insulation Sheet 1 Insulation Cover 1 TM-166 No 12 1 2 4 Q'ty Lower Case Assembly (1) Lower Case Assembly (1) Grounding Piece (A) SU B A Description 423- 3446 1 1 , A 7 6B91108A 8 6B6796 314-3601 . A Remarks 1 1 IB 3 O 5 ft 2 o 8 o 0 6 0 o O TITLE DRAWING NO. Lower Case Assembly (2) MODEL NO. ASSEMBLY NO. FX-2111 (CE) 422-3447 DATE 96.01.24 10-35 2 4-Double Semus Screw M3x& , G C 4 E F 1 D c D A G E 5 B O 3 F O 9 0 si fi' 6 /I 1 ii i" 41 V O Key Na Parts Na 1 9F3354 3 1M0381 Q ty Lower Case Assembly (2) 1 Recorder Assembly 1 684-2093 Thermal Array Head Cable 1 423-3444 PCB-6409 Assembly 1 422-3271 2 Description Drawing Na ' Remarks TITLE 4 5 5H9138 154-4577 Lead Label (C) 1 6 5H5639 154-3574 Lithium Battery Label 1 DRAWING NO. Lower Case Assembly (3) MODEL NO. ASSEMBLY NO. FX-2111 (CE) 422-3448 DATE 96.01.24 10-37 2 1 3§5 it r 5S> "J ir.> r re A z 3 Double Semus Screw, M3x8 B 4 A B si ' 4-Doubte t' Semus Screw M3x8 , Key Na Parts No. Drawing Na Description Q'ty Remarks TITLE 1 Lower Case Assembly (3) 1 2 Upper Case Assembly 1 Paper Magazine Assembly 1 Paper Shaft SU 1 3 9F2648 4 9H3522 224-0891 DRAWING NO. Exploded View MODEL NO. ASSEMBLY NO. FX-2111 (CE) 412-0928 DATE 96.01.24 10-39 FUKUDA DENSHI CO., LTD. 39-4, Hongo 3-chonne, Bunkyo-ku, Tokyo 113, Japan Phone:(03)3815-2121 Fax: (03)5684-3791