BV Family R1.2 / Volume 1
(04.0)
Copyright © 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS
RIGHTS RESERVED
RESERVED
SM – SW 0.5
PHILIPS
PHILIPS
SOFTWARE
INSTALLATION
1
BV-SCOPE
MENU ITEMS
2
DICOM
CONFIGURATION
3
PHILIPS MEDICAL SYSTEMS
S
ystem M
anual
System
Manual
S
oftware
Software
4
5
6
7
8
9
PMSN Best
Reproduction in whole or in part is prohibited without the written consent of
the copyright owner. Use of the information contained herein in any form
and/or by any means whatsoever is strictly reserved for Philips and its
licencees. Use of this manual by unauthorised persons is strictly prohibited.
10
Printed in Best, Netherlands.
BV Family R1.2 / Volume 1
(04.0)
Copyright © 2004
2004 Koninklijke Philips Electr
Electronics
onics N.V.
ALL RIGHTS
RIGHTS RESERVED
SM – SW 1
Service Manual BV Family R1.2
SERVICE MANUAL - SYSTEM:
SERVICE
SYSTEM: B V Famil y RELEA
RELEASE
SE 1.2 SOFTWARE
SOFTWARE
SYSTEM:
SYST
EM: BV Famil y R1.2
TYPE NO:
Manual Code Number 4522 981 22636
LIST OF PAGES and DRAWINGS
0.5
1
2.1
(04.0)
(04.0)
(05.0)
*
SM-SW 1-1 … 18
(04.2)
SM-SW 2-1 … 25
(05.0)
SM-SW 3-1 … 26
(04.1)
Date: 15-March-2005
BV Family R1.2 / Volume 1
(05.0)
Copyright © 2005 Koninklijke Philips Electro
Electronics
nics N.V.
ALL RIGHTS
RIGHTS RESERVED
SM – SW 2.1
Service Manual BV Family R1.2
Software installation
Sect
Se
ctii on 1
Section 1
SO
SOFT
FTWARE
WARE INS
INSTAL
TALLA
LATIO
TION
N
Contents
1
INTRODUCTION
INTRODUCTIO
N ........................................................................................
.....................................................................................................................................2
.............................................2
2
REQUIREMENTS OF THE SERVICE PC...............................................................................................2
3
CONNECTING THE SERVICE PC TO THE BV FAMILY SYSTEM.......................................................3
4
CONFIGURING THE CONNECTION
CONNECTION BETWEEN SERVICE PC AND BV FAMILY
FA MILY SYSTEM ..............5
4.1
Configuration of a connection with a serial cable ................................................................................
....................................................................................5
....5
4.2
Configuration of a connection with a cross cable or a network cable....................................................11
5
PREPARATION OF THE SERVICE PC FOR BV-SCOPE...................................................................12
5.1
Configuration of Internet Explorer settings ............................................................................................12
5.2
Configuration of ActiveX control settings...............................................................................................12
5.3
Installation of the BV-Scope hardware key............................................................................................14
5.4
Creating a short cut for BV-Scope.........................................................................................................14
6
DOWNLOADING THE SYSTEM SOFTWARE TO THE BV FAMILY SYSTEM ..................................16
7
PROCEDURES AFTER DFI REPLACEMENT OR SOFTWARE RE-INSTAL
RE-INSTALLA
LATION
TION .......................17
8
INSTALLA
INSTAL
LATION
TION OF E
EXTRA
XTRA SOFTWARE O
OPTIONS
PTIONS ..........................................................................18
8.1
Installation procedure ............................................................................................................................18
............................................................................................................................18
8.2
Check of new functions after installation of extra software options.......................................................18
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 1
Section 1
1
Service Manual BV Family R1.2
Software installation
INTRODUCTION
This sections deals with the confi guration o f a connection b etwee
etween
n the service PC and the BV Family
Family
syst em in case of BV-Scope
BV-Scope with integrated DI
DICO
COM
M functio ns (= performing DICOM functions without
external Merge Box).
NOTE
Systems with an external Merge Box have been delivered with another version of the system software.
These systems have to be configured conform the documentation delivered to those systems.
________________
The service PC and the BV Family system can be connected in 3 ways: with a serial cable, a cross cable and
a network cable. Depending on the way you choose, the service PC has to be configured and prepared. After
this, the system software for the BV Family system can be downloaded from the service PC to the BV Family
system.
CAUTION
Reinstallation / changing software is only allowed by authorized persons.
___________
______
__________
_____
WARNING
After adjustmen
adjustments
ts of the system always
always res
restart
tart the system to ensure,
ensure, the new
new adjustments
adjustments will be activated.
activated.
________________
WARNING
Together with a restart of the system, always close BV-scope and restart it.
________________
2
REQUIREMENTS OF THE SERVICE PC
The requirements for the service PC are:
1. Windows 2000 with administrator’
administrator’s
s rights.
2. Internet Explorer 5.5 or higher
3. CD-ROM player
4.
5.
6.
7.
8.
Ethernet-card (preferred) or serial port
A connected
connected hardware key (level
(level 2 fo
forr X-ray) with known password (see
(see chapter 5.3 if the hardware key
and its software has not already been installed).
PMSSec reader release 2.289 or higher.
Start PMSSec reader to check this.
(Usually this can be done by selecting: Programs > PMS Service Work Bench > PMSSec Reader . The
PMSSec Reader icon appears. Position the pointer on top of the icon. The release no. appears.)
Security toolbox release 1.16. or higher.
Start the Security toolbox to check this.
(Usually this can be done by selecting: Programs > PMS Service Work Bench > Security Toolbox.
Toolbox . The
Security Toolbox window appears. Select Help > About.
About. The version no. appears.)
In case of integrated DIC
DICOM
OM functionality:
functionality: contact
contact the network administrator for an IP address for the
service PC. This is not necessary if for service the service PC will be connected to the BV Family system
with a serial cable.
SM-SW 1 - 2
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
3
Section 1
CONNECTIN
CONNECTING
G THE SERVICE PC TO THE BV FAMIL
FAMILY
Y SYSTEM
The Service PC can be connected to the BV Family system in 3 ways: with a serial cable, a cross cable and a
network cable. See figure below. For the consequences, see the table below.
Connection with a serial cable (recommended connection for service)
Network
Ne
twork cable
BV Fam. System
(DFI in MVS)
Serial cable
(only for DICOM
functions)
DICOM Archive
DICOM Printer
Service PC
with
BV-Scope
interface
Connection
Conne
ction w ith a cross cable
DICOM Archive
BV Fam. System
(DFI in MVS)
Cross cable
Service PC
DICOM Printer
with
BV-Scope
interface
Connection
Conne
ction with a network cable
DICOM Archive
Network cable
BV Fam. System
(DFI in MVS)
DICOM Printer
Service PC
with
BV-Scope
interface
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 3
Section 1
Service Manual BV Family R1.2
Software installation
Possibl e connection s between the Se
Servic
rvic e PC and the BV Family
Family syst em
Cable
Connector
Serial cable,
see note 1
Cross cable,
cable,
see note 3
Network cable
Serial
Ac cess
Acces
s to
to B
BVVScope
possible?
Yes
Pe
Performing
rforming of
DICOM
DIC
OM function s
possible?
No, see note 2
Ethernet
Yes
Yes
Ethernet
Yes
Yes
Downloading of
Downloading
software
Type of
connection
Yes, takes about
Local
20 min.
Yes
(about 2 min.)
Yes
(about 2 min.)
(= point to point)
Local
(= point to point)
Network
NOTE 1
Connection w ith a serial cable is r ecommended, however see note 2.
___________
______
__________
_____
NOTE 2
When there is a connection with a serial cable, only service is possible. To perform DICOM functions a
network cable should be used to connect the BV Family system with the DICOM network.
________________
NOTE 3
With a only a cross cable, service of the BV Family system and performing of DICOM functions is not possible
at the same time (except when the DICOM application and BV-scope are running on the same PC).
________________
For the necessary configuration of the connection and the preparation of the service PC for BV-Scope, follow
the next chapters.
SM-SW 1 - 4
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
4
Section 1
CONFIGURIN
CONFIGURING
G THE CONNECTION BETWEEN SERVICE PC AND BV
FAMILY SYSTEM
The service PC can be connected to the BV Family system in 3 ways: with a serial cable, a cross cable and a
network cable. Configuration with a serial cable is recommended. For the limitations of this connection, refer
to the previous chapter.
4.1 Configuration of a connection with a serial cable p. 5
4.2 Configuration of a connection with a cross cable or a network cable p. 11
4.1
CONFIGURATION OF A CONNECTION WITH A SERIAL CABLE
This procedure also applies for reinstallation.
1. Connect a serial cable between the service PC and the BV Family system.
2. Select Start > Settings > Control Panel > Network and Dial-up Connections. The Network and Dial-up
Connections window appears (see figure below).
3.
Double click the Make New Connection icon. If you do this for the first time, the Location Information
window below appears. If this is not the first time, click the Next button and proceed with step 7 below.
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 5
Section 1
4.
5.
6.
Service Manual BV Family R1.2
Software installation
In the What country/region are you in now? field, enter ju st a cou nt
ntry
ry..
In the What area code (or city code) are you in now? field, enter ju st 4 dig it s .
Click the OK button. The Network Connection Wizard window appears.
7. Select Connect directly to another computer and click the Next button. The window below appears.
8. Select Guest and click the Next button. The window below appears.
SM-SW 1 - 6
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
9. Select Communications Port (COM1) and click the Next button. The window below appears
Section 1
9. Select For all users and click the Next button. The window below appears.
10. Enter a name for your connection.
11. Click the Finish button
button.. The window below appears.
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 7
Section 1
12. Change the User name field to: dfi
df i
13. Change the Password field to: software
14. Click the Properties button. The window below appears.
Service Manual BV Family R1.2
Software installation
15. Select Communications cable between two computer (COM1)
16. Click the Configure button. The Modem Configuration window appears.
SM-SW 1 - 8
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
17.
18.
19.
20.
In the Maximum speed field select 115200
Disable all hardware
hardware features.
Click the OK button. The window below appears.
Set up the Options, Security, Networking, and Sharing tabs as shown below:
Section 1
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 9
Section 1
Service Manual BV Family R1.2
Software installation
21. When set up as shown above, click
click the OK button. The Connect Direct Connection window appears.
22. To start the new connection enter
enter the password and click the Connect button.
To start the new connection from the Desktop, select Start > Settings > Control Panel > Network and Dial-up
Connections and select the connection with the name you entered in step 10 above.
If there are problems with the serial
serial co nnection:
nnection:
•
•
•
Restart MVS.
Enable Service on the MVS. (To enable service on the MVS refer to chapter 5.4 step 3.
3.))
Restart the Service PC.
SM-SW 1 - 10
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
4.2
Section 1
CONFIGURATION OF A CONNECTION WITH A CROSS CABLE OR A NETWORK CABLE
NOTE
This procedure is not necessary if the service PC will be connected to the BV Family system with a serial
cable.
________________
The settings for the network are described for a Windows 2000 operating system.
1. Start > Settings > Control Panel. The Control Panel window appears.
2. Double click the Network and Dail-up Connections icon. The corresponding window appears.
3. Click with the ri
right
ght m
mouse
ouse button on the Local Area Connection icon. A menu appears.
4. Click Properties
Properties.. The Local Area Connection Properties window appears.
5. Mark and select Internet Protocol (TCP/IP).
Click the Properties button. The Internet Protocol (TCP/IP) Properties window appears.
Select Use the following IP Address.
Click the Advance
Advanced
d button. The Advance
Advanced
d TCP/I
TCP/IP
P Settings
Settings window appears.
Click the upper Add button. The TCP/IP Address window appears.
Enter the IP address as shown in the Setup panel of the MVS. (If there is not shown an IP address: enter
the default IP address 192.168.1.1 )
11. Enter the Subnet mask:
mask: 255.255.255.0
12. Click the Add button
button.. The Advance
Advanced
d TCP/IP Settings
Settings window appears.
6.
7.
8.
9.
10.
13. Click OK. The Internet Protocol (TCP/IP) Properties window appears.
NOTE
If 'Use the following IP address' has been selected for BV-Scope to work and you use VRAS,
then select Obtain IP address automatically.
To use BV-Scope again, select Use the following IP address.
___________
______
__________
_____
14.
15.
16.
17.
Click OK. The Local Area Connection Properties window appears.
Click OK
OK.. The Network and Dail-up Connections window appears.
Close the window.
Restart the Service
Service PC.
NOTE
The Network card in the service PC has to communicate with the 10 MB Ethernet connection of the DFI.
Therefore, the communication speed of the network card has to be set at 10 Mbps or at Auto.
________________
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 11
Section 1
Service Manual BV Family R1.2
Software installation
5
PREPARATION OF THE SERVICE PC FOR BV-SCOPE
5.1
CONFIGURATION OF INTERNET EXPLORER SETTINGS
This procedure has been written for Windows 2000, for other Windows versions, the procedure may differ.
1.
Start Internet
2. Select
Tools >Explorer.
Internet Options.
Options. The corresponding window appears.
3. Select the General tab, click the Settings button.
4. Select Automat
Automatically
ically,, click the OK button.
5. Select the Security tab, click the Trusted sites icon.
6. Click the Sites button. The Trusted sites window appears.
7. Unselect Require server verification…
verification…
8. Start the mobile v
viewing
iewing sta
station
tion (MV
(MVS).
S). This is necessary to
to look up an IP
IP address.
9. In the Add this
this Web site
site to the zone field enter http:// followed by the IP address of the
the MVS (This IP
address is shown in the Setup panel of the MVS (the default IP address is 192.168.1.61
192.168.1.61).
). The Setup
panel can be reached by clicking the corresponding icon in the Patient Administration screen).
10. Click the Add button.
11. Click the OK button. The Security tab appears.
12. Select the Connections tab.
13. Select Never dial a connection.
connection.
14. Click the LAN Settings button. The Local Area Network (LAN) Settings window appears.
15. If Use a proxy server for your LAN has been selected, click the Advance
Advanced
d button. If not:
not : go to step 18.
16. In the Exception panel, enter the IP address of the MVS.
17. Click the OK button.
18. Close all windows
windows using the OK buttons. The entered values will be stored.
5.2
CONFIGURATION OF A CTIVEX CONTROL SETTINGS
ActiveX controls
controls must be enabled
enabled to readout the hardware
hardware key (dongle). Check the settings
settings as follows:
1. Start Internet Explorer.
2. Select Tools > Internet Options.
Options. The corresponding window appears.
3. Select the Security tab, the corresponding tab appears.
4. Click the Custom Level… button. The Security Settings window appears, see figure below.
SM-SW 1 - 12
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
5.
Check if the settings for the ActiveX controls
controls and pl
plug-ins
ug-ins are Enabled (see figures above):
Download signed ActiveX controls has to be enabled.
Download unsigned ActiveX controls has to be enabled.
Initialize and script ActiveX controls not marked as safe .. has to be enabled.
Run ActiveX controls and plug-ins has to be enabled.
Script ActiveX controls marked as safe for scripting has to be enabled.
−
−
−
−
−
6.
Scroll down and check if the Java permissions are set at Medium safety,
safety, see figure below.
Section 1
7.
8.
Click the OK button. The Security tab appears.
Click the OK button.
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 13
Section 1
5.3
Service Manual BV Family R1.2
Software installation
NSTALLATION
TION OF THE BV-SCOPE HARDWARE KEY
INSTALLA
A hardware key is
is required. Without
Without a hardware key the access
access is restricted.
restricted. The hardware key
key can only be
used for a limited period (one year). Upgrading is necessary after the one-year period has expired. Upgrading
can be done by the SSD/SSR service organization only.
For an initial installation of the hardware key software on the service PC refer to the Users Guide &
Installation Manual
Manual,, which is supplied together with the HW
HW-Key
-Key Toolbo x .
5.4
CREATING A SHORT CUT FOR BV-SCOPE
The function of this short cut is to give quick access to a certain IP address with a certain port number.
FirstConnect
t ime use
of BV-Scope
1.
a cable
between the
the servic
service
e PC and the MVS.
2. If PMSSec is active: in the
the Windows task bar right click the PMSSec icon (an icon with
with a horizontal key in
it). A menu appears. Click Close
Close..
WARNING
PMSSec must be closed otherwise, the file to be downloaded to the MVS will become unusable.
________________
3.
In the
the Setup
Setup panel of the MVS select Enable service. See figure below.
NOTE
If Enable service has not been selected in the Setup panel of the MVS, BV-Scope will not start and service is
not possible. This function has been added for security of patients’ data.
________________
SM-SW 1 - 14
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
4.
5.
6.
7.
8.
9.
10.
11.
12.
Section 1
Start Internet Explorer.
In the Addre
Address
ss bar, enter: http:// followed by the IP address shown
shown in the Setup panel of the MVS (the
default address is 192.168.1.61)
192.168.1.61) followed by: :54670
The File Download window appears.
Click the Save button. The Save As window appears.
In the Save as field, select Desktop
Desktop..
In the File name field change the name in: BV-Scope-DICOM.hta
Click the Save button. The Download Complete window appears.
Click Close
Close..
Close Internet Explorer.
NOTE
The BV-Scope shortcut gives quick access to BV Family systems with a certain IP address and a certain port
number. For BV Family systems with another IP address and/or port number a new shortcut may be created.
________________
13. Start BV-Scope
BV-Scope with the BV-Scope shortcut on the desktop. The Target Address Confirmation window
appears.
14. Enter the IP address shown
shown in the Setup panel of the MVS.
15. Click the OK button. The PMS Security Reader window appears.
16. Enter your hardware key password.
17. Click the OK button. BV-Scope will appear with its top-level menu.
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 15
Section 1
6
Service Manual BV Family R1.2
Software installation
DOWNLOADING THE SYSTEM SOFTWARE TO THE BV FAMIL
FAMILY
Y
SYSTEM
The system software for the BV Family system has to be downloaded from CD-Rom to the BV Family system
in case of a software update, a new release or reinstallation. The software can be downloaded in two ways:
With an Ethernet connection (with a network cable or cross cable). This takes about 2 minutes.
With a dial-up connection (serial cable). This takes about 20 minutes.
The two ways are described below. For a drawing of the connections, see SM-SW 11-3.
3.
•
•
NOTE
On a BV Family system the system software has already been installed in the factory, for installation of a new
DFI-2 skip steps 2, 4 and 5.
________________
1.
2.
3.
4.
5.
6.
7.
Turn on the MVS and wait until the system is
is operational.
operational.
In the
the Setup
Setup panel of the MV
MVS:
S: select enable service.
service.
Connect a cable between the service PC and the BV Family System:
a cross cable or network cable to the Ethernet port of the BV Family System or
a serial cable to the D-connector of the BV Family System.
In BV-Scope select Install > Program Manual > Network Configuration.
Enable the Extended Service Access.
Access.
Put the CD into the CD-ROM
CD-ROM drive
drive of the service PC that has been connected to the system.
Start the SWDownload.exe application on the CD-ROM. The window below appears.
In case of connection
connection with a cross cable or network cable:
cable: select Ethernet Connection.
In case of connection with a serial cable: select Dial-up Connection.
9. Enter the IP address shown on the Setup panel of the MVS.
MVS.
10. Click the Download Software button. The software will be downloaded to the DFI and the result will be
displayed in the output window.
8.
11. After about two minutes,
minutes, the message Download completed successfully appears: close the window and
switch the BV Family system off and on.
12. Close all windows.
13. The DFI will download the software to the Midas board in the C-stand if the software version is newer. It
will display the download progress on the User Interface of the C-stand. In addition, the SHA1 LEDs will
flash. The MVS will display being in “stand-alone” mode.
14. Wait until download will
will be finished (progress
(progress is 100% will be visible on the User Interface on the Cstand). The stand will reboot automatically.
15. After reboot, the complete system will be operational and standby.
standby.
SM-SW 1 - 16
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software installation
7
Section 1
PROCEDUR
PROCEDURES
ES AFTER DFI REPLA
REPLACEMENT
CEMENT OR SOFTWARE REINSTALLATION
Actions required after replacements of the DFI or a
after
fter re-installation of software.
1. Load software (see chapter 6)
2. Start BV-Scope.
3. Select Install > Program Automatic > Examination.
If you want to use the default examination types:
On the BV Family System, select Intern.
Click OK
OK..
Select a dataset valid for the system, see table below.
Click the OK button.
If you want to use client specific examination types:
On the Service PC, select Extern.
Browse to the file Exam-cu.lib for a Libra system
Exam-cu.end for an Endura system
Exam-cu.pul for a Pulsera system
Select the file and click the OK button.
Click the OK button.
−
−
−
−
−
−
−
−
Table
Table to select a dataset for a certain s ystem
System
Commercial optio ns
Type of Monitor
No of
Dataset
type
Examinations
(Exam file)
BV Libra
Basic
Standard / LCD
1 set
Standard
BV Libra
Basic
High-End
1 set
High-End
BV Libra
Basic, Vasc
Standard / LCD
2 sets
Standard
BV Libra
Basic, Vasc
High-End
2 sets
High-End
BV Endura
BV Endura
BV Endura
BV Endura
BV Endura
BV Endura
BV Pulsera
Basic
Basic
Basic, Vasc
Basic, Vasc
Basic, Vasc, CO2
Basic, Vasc, CO2
Basic
Standard / LCD
High-End
Standard / LCD
High-End
Standard / LCD
High-End
Standard / LCD
5 sets
5 sets
8 sets
8 sets
9 sets
9 sets
5 sets
Standard
High-End
Standard
High-End
Standard
High-End
Vasc/CO2/Cardio
Vasc/CO2/Cardio Standard
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
BV Pulsera
Basic
Basic, Vasc
Basic, Vasc
Basic, Vasc, CO2.
Basic, Vasc, CO2.
Basic, Cardio
Basic, Cardio
Basic, Cardio, Vasc
Basic, Cardio, Vasc
Basic, Cardio, Vasc,
CO2
Basic, Cardio, Vasc,
CO2
High-End
Standard / LCD
High-End
Standard / LCD
High-End
Standard / LCD
High-End
Standard / LCD
High-End
Standard / LCD
5 sets
8 sets
8 sets
9 sets
9 sets
8 sets
8 sets
11 sets
11 sets
12 sets
Vasc/CO2/Cardio
Vasc/CO2/Cardio High-End
Vasc/CO2/Cardio
Vasc/CO2/Cardio Standard
Vasc/CO2/Cardio High-End
Vasc/CO2/Cardio Standard
Vasc/CO2/Cardio High-End
Cardio/Vasc/CO2
Cardio/Vasc/CO2 Standard
Cardio/Vasc/CO2 High-End
Cardio/Vasc/CO2
Cardio/Vasc/CO2 Standard
Cardio/Vasc/CO2 High-End
Vasc/CO2/Cardio
Vasc/CO2/Cardio Standard
High-End
12 sets
Vasc/CO2/Cardio High-End
BV Pulsera
BV Family R1.2 / Volume 1
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 1 - 17
Section 1
Service Manual BV Family R1.2
Software installation
8
INSTAL
INSTALLA
LA TION OF EXTRA SOFTWARE OPTIONS
8.1
INSTALLATION
PROCEDURE
If a new SW license key has provided, follow the procedure below.
1. Select Install > Program Manual > System configuration.
2. In the SW license key field enter the new software string (provided by Philips).
3. In the Settings column, check the No. of examinations field (if applicable): for the required value see the
table above. Change the value if required.
4. Click the Apply button. A window appears. Read the text and continue accordingly.
5. Click the Info button. The installed options appear.
6. Check the installed options according
according to your software
software license key agreement.
agreement.
7.
8.
9.
10.
11.
8.2
Close
theOK
window.
Click the
button in BV-Scope window.
Close BV-Scope.
Restart the system.
Check the new functions, see
see below.
CHECK OF NEW FUNCTIONS AFTER INSTALLATION OF EXTRA SOFTWARE OPTIONS
Af ter i ns
nstal
tallat
latio
ion
n of
o f extr
ex tr a sof tw are opt io ns:
ns : perf
p erfor
orm
m tthe
he rel
relevant
evant ch
check
ecks
s on
o n ffun
unct
ct ions
io ns
Example for checks after installing the Vascular package
On the C-stand:
1. Choose Examination type.
type.
2. Check if the examination types Vascular abdominal,
abdominal, Vascular Cerebral and Vascular Peripheral are
available.
3. Choose one of these 3 new examinations types.
4. Check the flip mode functions:
Is it possible to select Subtract
Subtract,, Roadmap or Trace (black) ?
Choose Subtract
Subtract..
Check the functioning.
Choose Trace (black)
Check functioning.
Check if Roadmap appears on the C-arc display and on the monitor.
Check the Roadmap functions.
−
−
−
−
−
−
On the MVS:
1. Check on the monitor if the Pixel shift button and the Landmarking
Landmarking button can be used.
2. Use the buttons and check the functions.
Example for checks after installing of DICOM options
1. Perform network configuration.
2. Configure a printer or storage target.
3. Verify the DICOM export functions
functions (in other words
words:: check DICOM functions).
functions).
SM-SW 1 - 18
(04.2)
© 2004 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software functions
Section 2
Sect
Se
ctio
ion
n2
BVBV-S
Sco
cope
pe menu it
items
ems
Contents
1
TREE OF MENU & SUB MENU ITEMS..................................................................................................3
1.1
Authorization
Authorizatio
n level required
required and u
user
ser types ....................................................................................
............................................................................................5
........5
2
OVERVIEW OF
OF MAIN SOFTWARE
SOFTWARE FUNCTIONS .................................................................................6
3
BUTTONS................................................................................................................................................6
4
SUB MENU ITEMS UNDER INSTALL....................................................................................................7
4.1
/Access Control/System Password Configuration
Configuration ...................................................................................7
...................................................................................7
4.2
/Access Control/BV-Scope Password Configuration...............................................................................7
4.3
/Backup-Restore/
/Backup-Rest
ore/ ..........................................................................................
.....................................................................................................................................7
...........................................7
4.4
4.5
/Program Automatic/Configuration
Automatic/Configuration...............................................................................................
...........................................................................................................8
............8
/Program Automatic/Examination
Automatic/Examination .........................................................................................
............................................................................................................8
...................8
4.6
/Program Automatic/DICOM Targets.......................................................................................
Targets.......................................................................................................8
................8
4.7
/Program Automatic/Printer types............................................................................................................8
4.8
/Program Manual/System Configuration/.................................................................................................9
Configuration/.................................................................................................9
4.9
/Program Manual/User Configuration/........................................................................................
Configuration/ ...................................................................................................10
...........10
4.10
/Program Manual/Network Configura
Configuration/..........................................................................................11
tion/..........................................................................................11
4.11
/Program Manual/Energ
Manual/Energy
y Storage/ ....................................................................................................12
4.12
/Program Manual/X-ray
Manual/X-ray Tank/.........................................................................................
Tank/ ............................................................................................................12
...................12
4.13
/Program Manual/Examination
Manual/Examination Names/ .............................................................................................13
4.14
/Program Manual/Site Specific/..........................................................................................................13
4.15
/Program Manual/System Settings/....................................................................................................14
Settings/....................................................................................................14
4.16
4.16.1
4.16.2
4.16.3
/Program Manual/Reset
Manual/Reset System
System Data ..............................................................................................
...............................................................................................16
.16
/Program Manual/Reset
Manual/Reset System Data/Patient Image Database.......................................................
Database........................................................16
.16
/Program Manual/Reset
Manual/Reset Data/Reset Calibration Data/ ......................................................................17
/Program Manual/Reset
Manual/Reset Data/Reset
Data/Reset Examination Data/ ...................................................................17
4.17
/Program Manual/Demo Mode................................................................................................
Mode ...........................................................................................................17
...........17
4.18
4.18.1
4.18.2
4.18.3
4.18.4
4.18.5
/Program Manual/DICOM Target Configura
Configuration.............................................................................
tion.................................................................................17
....17
/Program Manual/DICOM target Configuration/Export
Configuration/Export Targets .........................................................17
/Program Manual/DICOM Target Configuratio
Configuration/Printer
n/Printer Targets........................................................18
/Program Manual/DICOM Target Configuration/Worklist Query Definition........................................21
/Program Manual/DICOM Target Configuratio
Configuration/Worklist
n/Worklist Management.............................................22
/Program Manual/DICOM Target Configuration/Modality
Configuration/Modality Performed Procedure Step ......................23
4.19
/Program Manual/DICOM Queue management......................................................
management.................................................................................23
...........................23
4.20
4.21
/Adjustment ....................................................................................
........................................................................................................................................24
....................................................24
/Performance
/Performa
nce Test..............................................................................................................................24
Test..............................................................................................................................24
BV Family R1.2 / Volume 1
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 2 - 1
Section 2
Service Manual BV Family R1.2
Software functions
4.22
/Copy/Examination
/Copy/Examin
ation ..................................................................................................
.............................................................................................................................24
...........................24
4.23
/Copy/DICOM Targets........................................................................................................................24
Targets........................................................................................................................24
5
SUB MENU ITEMS UNDER FAULTFIND.............................................................................................25
SM-SW 2 - 2
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software functions
1
Section 2
TREE OF MENU & SUB MENU ITEMS
Software tree of BV-Scope for Service users
INSTALL
ACCESS CONTROL
System Password Configuration
BV Scope Password Configuratio
Configuration
n
BACKUP/RESTORE
PROGRAM AUTOMATIC
FAULTFIND
LOGGING
Logfile Complete System
Energy Storage history (Ra only)
Tank history
System history and calibration
SERVICE UTILITIES
Configuration
Examination
DICOM Targets
Printer types
PROGRAM MANUAL
System Configuration
User Configuration
Network Configuration
Energy Storage (RA only)
X-ray Tank
Examination Names (BV Endura and BV Pul sera only)
Site Specific
System Settings (with sub items)
Reset System Data (with sub items)
Demo Mode
DICOM target Configuration (with sub items)
DICOM queu management
ADJUSTMENT
Retrieve error list
Retrieve DFI error list
Retrieve Suco error list
X-ray tube
Fluoroscopy (FA only)
Conditioning
Filament
Dose output
Dose limitation
II/TV
Mechanical
Basic lens focussing (12" only)
Optical camera focus
Centre of rotation
Centering image
Camera rotation
Electrical
Electrical II Focus
Video levels
Dose rate
Image Processor
Monitor image
Measure function
Collimator
Iris
Shutter A
Shutter B
Laser alignment
PERFORMANCE TEST
Image Quality
Level 1
Lower Level
Indicator / Buzzer
Key test
X-Ray tube performance
COPY
Examination
DICOM Targets
BV Family R1.2 / Volume 1
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 2 - 3
Section 2
Service Manual BV Family R1.2
Software functions
Example of a BV-Scope window for service users (authorization levels S0, S1, S2 and S3).
Remark: a grayed out item is not selectable.
Example of a BV-Scope window for BV-Scope administrators (authorization level A).
SM-SW 2 - 4
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software functions
1.1
Section 2
A UTHORIZATION LEVEL REQUIRED AND USER TYPES
BV-Scope uses a hardware key (dongle) to protect serviceable functions. As applied within Surgery, it is only
allowed to be used by authorized users.
To be able to offer the relevant functionality to the categories of BV-Scope users, a protection mechanism is
defined which matches the possible protection levels. There are 3 levels of authorization, see table below.
Au thor
th orizat
izatio
ion
n level
l evel r equi red an
and
d us er ttyp
ypes
es
Au th orizati
Auth
or izati on l evel
User types (levels)
(levels)
Passw
Passw ord required
PM
PMS
S sec & Dongle
required
X*
BV-Scope
BV-S
cope Admini strator (A)
Dongle level 0, 1 or 2
BV-Scope Servi ce (S0, S1, S2)
S2)
Dongle level 3
BV-Scope Factory (S3)
*) This password can be configured by user types with the levels S0, S1, S2 or S3).
In this definition, level S3 means the highest access
access ri ghts
ghts.. In practice, this means that these procedures are
allowed to be used by the smallest category of users (e.g. high level specialists). In special occasions, it can
be useful to make use of authorization level S3: contact the helpdesk.
The BV-Scope administrator (level A
A),
), has the lowest access rights. The BV-Scope administrator does not
require a hardware key, but has restricted access.
For all user types, only the permitted procedures are visible.
(The information above is a copy of the information in SM-G section 2.)
BV Family R1.2 / Volume 1
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 2 - 5
Section 2
2
What to do
Install the system
Customizing
Maintenance
IQ performance
Access Control
Control
3
Service Manual BV Family R1.2
Software functions
OVERVIEW OF MAIN SOFTWARE FUNCTIONS
Menu item
Menu
Install > Backup - Restore
Install > Program Automatic
Install > Program Manual
Install > Adjustment
Install > Copy
Install > Program Manual
Install > Adjustment
Install > Performance Test
Install > Adjustment
Faultfind > Logging
Install > Performance Test
Install > Access Control
BUTTONS
The following buttons are used in BV-Scope:
Intended use
Connecting two subsystems
Loading Config. Data, Exam. Data.
Customizing the system
Adjustment
Adjustm
ent and verification
verification of the
the system
Backup Exam.
Customizing the System data
Adjustment
Adjustm
ent and verification
verification of the
the system
Checking the system
Adjustment
Adjustm
ent and verification
verification of the
the system
Faultfinding
Checking the performance of the system
Changing the system and BV-Scope
password
Button
Function
next
repeat
Go to next screen without storage of
values
Go to previous screen without storage of
values
Repeat current step of adjustment
ok
Store values and go to next screen
apply
Store values
cancel
Go to main BV-scope menu without apply
of values of current step
Go to help info
previous
help
info
save
exit
check
SM-SW 2 - 6
Note: read the instruction window in BV–
Scope for the correct functioning of the
OK button.
Note: clicking the apply button is ignored
during radiation
Displays info about software packages
installed
Save current page
Exit BV-Scope application
Test values
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software functions
4
Section 2
SUB MENU ITEMS UNDER INSTAL
INSTALL
L
For the default settings of values, see the relevant Release Bulletins.
CCESS
4.1
/A
ONTROL
C
YSTEM
/S
ASSWORD
P
ONFIGURATION
C
This menu item is optional. Your current software version may not contain this option.
Settings (authorization level required ≥ A)
Configuration item
Password
4.2
Value
Value set
string
Default
n.a.
Explanation
Password of the user of the system
/A CCESS CONTROL/BV-SCOPE PASSWORD CONFIGURATION
Settings (authorization level required ≥ A)
Configuration item
Value
Value set
Default
Explanation
Administrator
Adminis
trator name
string
n.a.
Name o
off the administr
administrator
ator of
of the system
system
Password
4.3
string
n.a.
Password of the administrator of the system
/B ACK
ACKUP
UP-RESTORE/
Reference
Configuration item
System Serial Number
Value
Value set
xxx
Explanation
Serial no. of the system filled by the factory
Settings (authorization level required ≥ S2)
Configuration item
Action
Mobile Viewing Station serial
number
X-ray Control serial number
BV Family R1.2 / Volume 1
Value
Value set
No actio
action
n
Explanation
-
Reconnect
Service connection
Used after replacement of Midas or DFI.
Used to make a test connection for faultfinding
purposes between two usually not connected
subsystems (MVS and CAS)
Serial number of Mobile Viewing Station
Enter text string
Enter text string
Serial number of X-ray Control (on the C-arm
Stand)
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 2 - 7
Section 2
4.4
Service Manual BV Family R1.2
Software functions
/PROGRAM A UTOMATIC/CONFIGURATION
Settings (authorization level required ≥ S3)
Configuration item
Value
Value set
Explanation
Configuration File
Intern (BV Family system)
Extern (Service PC)
Select Intern
tothe
load
the Configuration File from the
internal
disk to
system.
Select Extern to load the Configuration File from the
diskette of the Service PC to the system.
File
File name
In the File field, the name of a Configuration File has
to be entered. With the Browse button, this file can
also be selected.
4.5
/PROGRAM A UTOMATIC/EXAMINATION
Settings (authorization level required ≥ S0)
Configuration item
Select Examination File
Value
Value set
Intern (BV Family system)
Explanation
Select Intern to load the default examination data.
Select Extern to load the examination data from the
Extern (Service PC)
4.6
Service PC to the system.
/PROGRAM A UTOMATIC/DICOM TARGETS
(Authorization level required ≥ A)
In the File field, the name of the data file of a DICOM Target has to be entered. With the Browse button, this
file can also be selected. With this function, a back up of the DICOM configuration made with the function
‘Install/Copy/DICOM Targets’ can be restored.
4.7
/PROGRAM A UTOMATIC/PRINTER TYPES
≥
(Authorization
required
A)
In the File field,level
the name
of pre-configures
Printer Types file has to be entered. With the Browse button, this
file can also be selected. The Printer Types File (Dicom.ptf) can be found on the BV Family Software cd-rom,
in the subdirectory: dfi\data.
SM-SW 2 - 8
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
BV Family R1.2 / Volume 1
Service Manual BV Family R1.2
Software functions
4.8
Section 2
ANUAL
L /SYSTEM CONFIGURATION/
/PROGRAM MANUA
Reference
Configuration item
System type:
X-ray Tank installed
Battery pack installed
Value
Value set
BV Endura-9”, BV Endura-12”
BV Pulsera-9”, BV Pulsera-12”
BV Libra-6”, BV Libra-9”
dd-mmm-yyyy
dd-mmm-yyyy
SW release
Explanation
Type of BV Family system
Installation date of the X-ray Tank
Installation date of the Energy Storage Unit
SW version number
Hospital Name
Read from setting
Name set in the setting window
Host Identification
Video Frequency [Hz]
Read from setting
50
8-digit Identification number
This setting cannot be changed. The setting is
independent of the mains frequency (50/60 Hz).
Settings (authorization level required ≥ S2)
Configuration item
Value
Value set
Explanation
System Serial number
text string
Serial number of the system
Shutter L + R Coupled
Yes, No
Auto Shutter
Shutter Position
Positioning
ing
Enab
Enable,
le, Disable
Disable
Camera angle [deg]
0 .. 359
Radiography possible
Yes, No
possible.
Warning: Ra
Radiography
diography with 6" BV Libra is not permitted
permitted
in t he USA.
If No has been selected: only fluoroscopy is possible.
Selection depends on mains supply conditions and/or legal
requirements.
Maximum number of examinations depending on system
type and options. To display the options use the info button
No. of examinations
types
1 - 13
Max kV
40…max kV
[kV]
French Homologation
Yes, No
Gamma Correction
1…4
Initialize Image Disk
Yes, No
SW License key
Text string
BV Family R1.2 / Volume 1
Coupling of movements of the shutters
Set
Set upon u ser's preference
Enable
Enable or d
disabled
isabled the automatic
automatic position
positioning
ing of the shutters
shutters
of the collimator.
Default position of the camera angle.
Set
Set upon u ser's preference
If Yes has been selected: radiography and fluoroscopy are
in the upper frame of the window.
Maximum voltage of the X-ray tube
Depends on legal requirements
French Homologation can only be selected on FA systems.
Depends on legal requirements
Caution: after French Homologation has been enabled, it
must be calibrated with Filament (tube) adjustment.
Always select “1”
All images are removed if Yes is selected; no other data is
removed. Selection of Yes is necessary when the max
number of images to be stored is changed.
License string.
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
SM-SW 2 - 9
ALL RIGHTS RESERVED
RESERVED
Section 2
4.9
Service Manual BV Family R1.2
Software functions
/PROGRAM MANUA
ANUAL
L /USER CONFIGURATION/
Reference
Configuration item
Value
Value set
SW release
Host Identification
Explanation
SW version number
Read from setting
8-digit Identification number
Settings (authorization level required ≥ A)
Configuration item
Value
Value set
Explanation
Hospital name
Language
Text string
English, French,
German, Spanish,
Swedish
Birth date, Remark
Set
Set upon user's preference
Default language of the User Interface on the stand and the
Mobile Viewing Station.
Set
Set upon user's preference
Field can be set for fixed date format or as free remark
Set
Set upon user's preference
Date format
Set
Set upon user's preference
Birth date field as
Data Format
dd-mm-yyyy,
yyyy-mm-dd,
Live additional text
mm dd yyyy
On, Off
Display Exam type
Units of measure
Yes, No
0.1 mm, 0.01 inch
SM-SW 2 - 10
Hospital name, name of physician, examination type,
examination date and time Yes/No visible during live fluoro
Set
Set upon user's preference
Set
Set upon user's preference
Measure function in inch or mm
Set
Set upon user's pr eference
eference
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
BV Family R1.2 / Volume 1
ALL RIGHTS RESERVED
RESERVED
Service Manual BV Family R1.2
Software functions
4.10
Section 2
ANUAL
L /NETWORK CONFIGURATION/
/PROGRAM MANUA
Reference
Configuration item
Value
Value
Explanation
SW Release
Ethernet address
RR.VV.LL
XX.XX.XX.XX.XX.XX
Release number. Version number. Level number.
Unique ID no. of the network card of the BV Fam.
System.
Also called
called MAC address
address
Settings
Configuration
item
IP address
Subnet mask
Default Gateway
Extended Service
Access
Station name
Value set/range
Default
De
fault
Explanation
Explanation
###.###.###.###
###.###.###.###
###.###.###.###
Checkbox (Enabled,
Disabled)
20 characters, alpha-
Read from setting
Read from setting
Read from setting
Disabled
IP address of BV Family system
Subnet mask of BV Family system
Gateway used by BV Fam. system
Enables access.
Required for software download.
DICOM station name of the BV
“No name”
Auth
level
≥ A
≥ A
≥ A
≥ S2
≥
A
numeric
DICOM Configuratio n
DICOM
AE Title
20 cha
characters
racters,, alphaalphanumeric
Interpolation
On, Off
Family system
“No name”
DICOM Application Entity title of the
BV Family system
If images are not square and targets
cannot handle this, it is needed to
export images with interpolation on.
Interpolation determines the aspect
On
≥
A
≥
A
ratio
of the pixels
of the image.
Interpolation
On: aspect
On:
ratio: 1 : 1
Interpolation Off:
aspect ratio: 549 : 753
NOTE
The BV Family uses port 8104 for DICOM Storage Commit and port 104 for all other DICOM AEs.
___________
______
__________
_____
Remark: the parameters below can be used by the IT specialist of the hospital network to optimize the system
and/or network load and the target performance.
Configuration item
Value
Value set/range
set/range
Maximum
(PDU) SizeProtocol Data Units
Recei
Rec
eive
ve me
messa
ssage
ge Ti
Time
meou
outt
Association
Associatio
n close timeout
Association
Associati
on rep
reply
ly timeout
timeout
Association
Associati
on rele
release
ase timeout
timeout
Write
Wr
ite tim
timeo
eout
ut
Networ
Net
work
k Con
Conne
nect
ct Tim
Timeo
eout
ut
4….256 kB
1…
1…36
3600
00 se
sec
c
1…3600
1…3600 sec
1…360
1…3600
0 sec
1…360
1…3600
0 sec
1…36
1…3600
00 se
sec
c
1…36
1…3600
00 se
sec
c
60
1
60
60
12
120
0
60
≥
Network
Netw
ork Inactivi
Inactivity
ty –Timeo
–Timeout
ut
1…3600
1…3600 sec
60
≥
BV Family R1.2 / Volume 1
Default
fault
Expl
xpla
ana
natio
tion
n
28
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
Auth.
Auth.
level
≥
A
A
≥ A
≥ A
≥ A
≥ A
≥ A
A
SM-SW 2 - 11
ALL RIGHTS RESERVED
RESERVED
Section 2
Service Manual BV Family R1.2
Software functions
Digital Navigation Link Option
Configuration item
Value
Value set/range
set/range
Default
Explanation
Port number
1025 - 65555
55555
Port number of the digital navigation
link port.
Change the port number in case it is
requested by the FSE from the
company that delivers the
navigational tool.
4.11
/PROGRAM MANUA
ANUAL
L /ENERGY STORAGE/
Reference
Configuration item
Prev. installation:
Value
Value set
Date
Settings (authorization level required ≥ S2)
Explanation
Previous installation date of battery-pack
≥
A
Configuration item
Install new Battery-pack
4.12
Value
Value set
Yes, No
Explanation
Selection of Yes will store the current date in the
field 'Battery-pack installed' under /Program
Manual/System Configuration
/PROGRAM MANUA
ANUAL
L /X-RAY TANK /
Reference
Configuration item
Prev. installation
Explanation
Previous installation date of X-ray tank
Value
Value set
Date
Settings (authorization level required ≥ S2)
Configuration item
Install new tank
SM-SW 2 - 12
Value
Value set
Yes, No
Explanation
Selection of Yes will store the current date in the
field 'X-ray tank installed' under /Program
Manual/System Configuration
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
BV Family R1.2 / Volume 1
ALL RIGHTS RESERVED
RESERVED
Service Manual BV Family R1.2
Software functions
4.13
Section 2
ANUAL
L /EXAMINATION NAMES/
/PROGRAM MANUA
Reference
Configuration item
Value
Value set
Explanation
Last Exam. update
Date
Date of the last examination update
Settings (authorization level required ≥ A)
Configuration item
XX Examination name
06 - etc.
Value
Value set
'Name of examination'
Explanation
The examination names 01 up to 05 are
predefined, cannot be changed and are not
visible. The examination name will be used on
the user interface of the stand.
4.14
ANUAL
L /SITE SPECIFIC/
/PROGRAM MANUA
Settings (authorization level required ≥ S0)
Configuration item
Initialize Image Disk
Value
Value set
Yes, No
Run Buffer Size
10…5000
Cocktail [1] Contrast*
50 .. 150
Cocktail [2] Contrast*
50 .. 150
Cocktail [3] Contrast*
50 .. 150
Cocktail [1] Brightness*
50 .. 150
Cocktail [2] Brightness*
50 .. 150
Cocktail [3] Brightness*
50 .. 150
Explanation
In this screen, select Yes when you have changed the
run buffer. Ca
Caution:
ution: all images
images w ill be los t.
Gives the amount of images that is always left free for
storing runs.
Default value = 115 %
Can eventually be set
Default value = 130 %
Can eventually be set
Default value = 150 %
Can eventually be set
Default value = 104 %
Can eventually be set
Default value = 109 %
Can eventually be set
Default value = 113 %
Can eventually be set
*) Setting of a cocktail of the Contrast/Brightness keys on the stand and the Mobile Viewing Station.
BV Family R1.2 / Volume 1
(05.0)
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
SM-SW 2 - 13
ALL RIGHTS RESERVED
RESERVED
Section 2
4.15
Service Manual BV Family R1.2
Software functions
ANUAL
L /SYSTEM SETTINGS/
/PROGRAM MANUA
The following predefined system settings can be chosen:
•
Examination type
•
X-mode
•
•
Flip-mode
NR-level (BV Libra only)
These predefined settings can be modified with the settings mentioned below.
Reference
Configuration item
Actual Examina
Examination
tion
Value
Value set
Name of
of the actual
actual examination
examination..
Explanation
Last Examination update
Date
Date of last change of settings.
Settings (authorization level required ≥ S0)
(to make changes to the predefined settings above)
Configuration item
Common:
Value
Value set
Explanation
kV/mA curve
Normal, High Contrast,
Isowatt
Focus
Small, Large
XTV Measure Field Size
0 = auto, 1 ... 7
High contrast and Isowatt kV/mA curve are selections
for Pulsera only. Caution: For t he HQ Ortho
Ortho and
Thorax Examination
Examination c hanging kV/mA cur ve can
result in unexpected kV stabilized values.
Select small or large X-ray focus. The X modes
Pulsed and Digital can only be used with a large focus
Relative field size, fixed size 20% up to 70% or
automatic mode.
Leave in automatic mode (only 20% is feasible)
Left button
Image processing:
Enable subtraction
Yes, No
Enables / disables subtraction parameters
Trace black/white
Black, White
Depends on the commercial option, trace white (CO2)
or trace black. Depends on used contrast medium (For
iodine select Black, for CO2 select White)
Landmarking level
Noise reduction level
Image storage:
Storage Rate [fr/s]
0 ... 3
1…6
Default landmark level during subtraction
Default noise reduction level number (Libra only)
0, 1, 2, 3, 5, 8, 12, 25
Store Last Image Hold
Yes, No
VCR Control
On, Off
Default setting of image store rate, 0, 1, 2, 3, 5, 8, 12,
25 depending on disk capacity. In case of pulsed
exposure, the frame rate is equal to the pulse rate.
Set
Set upon user's pr eference
eference
If enabled, Last Image Hold will be stored
Set
Set upon user's pr eference
eference
Automatic switch on the video cassette recorder or
Medical DVD-Recorder: yes or no.
Set
Set upon user's pr eference
eference
SM-SW 2 - 14
(05.0)
BV Family R1.2 / Volume 1
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
Service Manual BV Family R1.2
Software functions
Section 2
Video
Vide
o processing:
Integration factor
0 .. 60
Required number of image frames in order to achieve
a desired noise reduction level. Change the value
only after consulting an application
application s pecialist
pecialist
FM Noise Reduction Level
1, ¾, ½, ¼, 1/8, 1/16
K-factor, specifies the level of noise reduction to be
(FM = Fluoro Mode)
applied
during fluoroscopy.
Change
the list
value only
afterr c onsulting
afte
an application
applica
tion specialist
specia
FM Movement Detection
On, Off
Movement switches back Noise reduction per pixel or
when movement is detected during fluoroscopy
FM Movement Sensitivity
Off, low, medium, high
Do not change this v alue
FM Contrast
-49 … + 49
Default position of Contrast ruler in Mobile Viewing
Station Post proc screen during fluoroscopy
FM Brightness
-49 … + 49
Default position of Brightness ruler in MVS Post proc
screen during fluoroscopy
FM Edge Enhancement
Off, 1 .. 15
Default position of Edge ruler in MVS Post proc
screen during fluoroscopy
FM Video Invert
On, Off
Default 'position' of Invert button on MVS user
interface during fluoroscopy
SM Noise Reduction Level
1, ¾, ½, ¼, 1/8, 1/16
K-factor, specifies the level of noise reduction to be
applied during subtraction. Change the value
value onl y
afterr c onsulting an application
afte
application specialist
specialist
SM Movement Detection
On, Off
Movement switches Noise reduction locally or when
movement is detected during subtraction
SM Movement Sensitivity
Off, low, medium, high
Do not change this v alue
SM Contrast
-49 … + 49
Default position of Contrast ruler in MVS Post proc
screen during subtraction
SM Brightness
-49 … + 49
Default position of Brightness ruler in MVS Post proc
screen during subtraction
SM Edge Enhancement
off, 1 - 15
Default position of Edge ruler in MVS Post proc
screen during subtraction
SM Video Invert
On, Off
Default 'position' of Invert button on MVS user
interface during subtraction
(SM = Subtraction Mode)
Right button
X-ray generator
Pulse width
Initial pulse rate
Image processing:
Enable subtraction
Trace black/white
Landmarking level
BV Family R1.2 / Volume 1
Short, Medium, Long
3, 5, 8, 12, 25
Duration of the pulse (Pulsera only)
Initial pulse rate in frames/sec (Pulsera only)
Yes, No
Black, White
Enable or disable the subtraction function
Depends on the commercial option, trace white (CO2)
or trace black. Depends on used contrast medium (For
iodine select Black, for CO2 select White)
Default landmark level during subtraction.
0 .. 3
(05.0)
SM-SW 2 - 15
© 2005 Koninklijke Philips Elect
Electronics
ronics N.V.
ALL RIGHTS RESERVED
RESERVED
Section 2
Service Manual BV Family R1.2
Software functions
Image storage:
Storage Rate [fr/s]
0, 1, 2, 3, 5, 8, 12, 25
Store Last Image Hold
Yes, No
VCR Control
On, Off
Default setting of image store rate, 0, 1, 2, 3, 5, 8, 12,
25 depending on disk capacity. In case of pulsed
exposure, the frame rate is equal to the pulse rate.
Set
Set upon user's pr eference
eference
If enabled, Last Image Hold will be stored
Set
Set upon user's pr eference
eference
Automatic switch on the videocassette recorder: Off or
On. Se
Sett upon u ser's preference
Video
Vide
o processing:
Integration factor
0 … 60
Required number of image frames in order to achieve
a desired noise reduction level
FM Noise Reduction Level
1, ¾, ½, ¼, 1/8, 1/16
K-factor, specifies the level of noise reduction to be
applied during fluoroscopy
FM Movement Detection
On, Off
Movement switches back Noise reduction per pixel or
when movement is detected during fluoroscopy
FM Movement Sensitivity
Off, low, medium, high
Do not change this v alue
FM Contrast
-49 … + 49
Default position of Contrast ruler in MVS Post proc
screen during fluoroscopy
FM Brightness
-49 … + 49
Default position of Brightness ruler in MVS Post proc
Download