Uploaded by 123456 789

Sterilization in dentistry

advertisement
STERILIZATION OF INSTRUMENTS
Sterilization - the complete elimination or destruction of all forms of
microorganisms → bacteria, viruses, fungi and spores.
High-level disinfection - destroys all microorganisms, except high levels of
bacterial spores.
The difference between HLD and sterilization is that only sterilization can
kill high numbers of bacterial spores.
Intermediate-level disinfection – kills Mycobacterium tuberculosis,
vegetative bacteria, most viruses, and most fungi, but it does not necessarily
kill bacterial spores.
Low-level disinfection - kills most bacteria, some viruses and some fungi,
but it cannot be relied on to kill resistant microorganisms such as tubercle
bacilli or bacterial spores.
It is impossible to sterilize all instruments, items, and surfaces etc that
become contaminated during dental procedures.
The choice of decontamination regimes may be based on how an item
or instrument will be used.
Instruments are defined as critical, semi-critical and non-critical.
(CDC)
Critical instruments: scissors, forceps, elevator and scaler. If an
instrument will be used to penetrate tissue or to touch bone it must be
sterilized.
Semi-critical instruments: mirror, probe, amalgam carrier and
tweezers. If an instrument will touch mucous membranes, but will not be
used to penetrate tissue or to touch bone, it should be sterilized if al all
possible or, if the instrument is susceptible to heat damage, it should be
subjected to high-level disinfection.
Non-critical instruments: These are defined as equipments and
surfaces which contact only intact skin, such as mixing slabs and spatulas.
Work surfaces are decontaminated by using intermediate-level disinfection.
Summary:
- Heat sterilizes all critical instruments and semi-critical instruments
which are not damaged by heat.
- Semi-critical instruments damaged by heat may be treated after use by
high-level disinfection
- Non-critical environmental surfaces are decontaminated using
intermediate-level disinfection.
CATEGORIES OF TASKS, WORK AREAS AND PERSONNEL
The following guidelines evaluate and classify tasks undertaken in a
dental practice into one of three categories.
Category I: tasks that involve exposure to blood, body fluids or
tissues.
- All procedures or other job-related tasks that involve an inherent
potential for contact of mucous membrane or skin with blood, body
fluids, or tissues, or a potential spills or splashes of these.
- Every employee engaged in category I tasks should be required to use
appropriate protective measures.
- Most (but not all!) tasks performed by the dentist, oral hygienist,
chairside assistant and laboratory technician would fall into this
category.
Category II: tasks that involve no exposure to blood, body fluids or
tissues, but personnel carrying out these tasks may be required to
perform unplanned Category I tasks.
- The normal work routine of these personnel involves no exposure
to blood, body fluids or tissues, but the understanding that
exposure or potential exposure may be required is a condition of
employment.
- Every employee engaged in category II tasks should have
appropriate protective measures readily available.
- Tasks performed by clerical or non-profesional workers who may,
as part of their duties, help to clean up the office, handle
instruments or impression materials, or send out dental materials to
laboratories, would be classified as Category II.
Category III: tasks that involve no exposure to blood, body fluids or
tissues.
- The normal work routine of these personnel involves no exposure
to blood, body fluids or tissues. They are not called upon , as part
of their employment, to perform or assist in emergency medical
care or first aid, or to be potentially exposed in some other way.
- A front-office receptionist, book-keeper, or insurance clerk who
does not handle dental instruments or materials would be a
Category III worker.
These classifications are not rigid and there may be crossover, depending on
the job performed.
Critical and semi-critical items and instruments are sterilized by heat.
There are four distinct stages which achieve safe instrument sterilization:
 Pre-cleaning disinfection, using “holding” solutions.
 Pre-sterilization cleaning
 Sterilization
 Aseptic storage.
The sterilization area
The layout of the sterilization area
Dirty
Sink Cleaning Ultrasonic
area
area
bath
Area of high contamination
Packaging
Sterilizer Clean
area
area
Area of
Area of low
medium
contamination
contamination
The area for cleaning and sterilization needs careful planning with a
generous amount of room allowed for wide worktops, a sink, an ultrasonic
cleaner and sterilizer(s).
Some offices have a separate central sterilization area. This should be
easily accessible to all the separate surgeries in the practice. Instruments
must be carried safely to surgery areas; closed trays, baskets, or trolleys are
recommended.
Many offices locate the sterilization area within the surgery. If this is
the case, it should be situated away from the operating area.
Dirty area
Trays containing contaminated instruments and disposables are taken
to this area after use. Disposables are placed carefully in a waste receiver.
Sink
A deep-sink with elbow or foot operated tap controls and an efficient
splash back is essential. A deep sink is necessary to minimize splashing
when washing or rinsing instruments.
The sink is for cleaning only. A separate sink, located nearby, should
be reserved for hand washing. Decontamination of rubber utility gloves may
be carried out in the cleaning sink. Instruments in holding baths are rinsed
and then taken to the cleaning area.
Cleaning area
After removal of contaminated disposable items, instruments and
trays are taken to the cleaning area, where they are organized into baskets
prior to ultrasonic cleaning. If trays cannot be ultrasonically cleaned, they
are wiped clean in this area, which should be sufficiently large for several
trays.
Ultrasonic bath
It should be a good quality product which cleans large number of
widely spaced instruments.
Packaging area
After cleaning, instruments are thoroughly rinsed and taken to the
packaging area where they are dried and then either loaded into trays, placed
in pouches, or wrapped. Trays are wrapped in this area.
Sterilization pouches, paper wraps, chemical indicator strips are stored
in cupboards above the packaging area.
Sterilizer
- A good quality autoclave
- A second sterilizer (a chemiclave) for use both as back-up and to
sterilize instruments which would be damaged by autoclaving.
Clean area
Hot trays and packaged instruments are taken from the sterilizer to
this area before being stored in adjacent closed cabinets.
Precautions
Adequate lighting is essential to facilitate careful inspection and
handling of instruments.
Heavy rubber utility gloves, protective eyewear, a mask and a plastic
apron should be worn when performing the decontamination stages of
sterilization.
PRE-STERILISATION DISINFECTION
After use, place the instruments into a disinfectant detergent solution
in a container located within the operating zone near to the dentist.
At the end of the dental procedure, take the container to the
sterilization area. Thoroughly rinse the instruments with water. The solution
is reused but should be discarded daily.
Place heavily contaminated endodontic files and rotary instruments
into a small volume of holding solution contained in a beaker located on the
bracket table tray. Discard holding solutions, rinse the instruments and add
ultrasonic solution to the beaker, then suspended this in an ultrasonic bath.
PRE-STERILIZATION CLEANING
Proteinaceous material protects microorganisms on the surface of
instruments from heat and other sterilization conditions. Pre-sterilization
cleaning may be achieved in one of three ways:
1. Hand scrubbing of instruments
2. Ultrasonic cleaning
3. Dishwasher instrument cleaning
Heavy rubber utility gloves, protective eyewear, a mask and a plastic
apron should be worn when decontaminating instruments in the
sterilization area.
1. Hand scrubbing of instruments
Hand scrubbing increases the chance of injury from contaminated
instruments and from contact with contaminated fluids. This method
of cleaning should be avoided if at all possible.
If an item must be manually scrubbed, splatter should be kept to a
minimum. This is achieved by scrubbing under water in a deep sink,
not under fast-running water from a tap. A long-handled brush is
always used.
2. Ultrasonic cleaning → more efficient than manual cleaning.
When correctly performed, ultrasonic cleaning will remove dried
serum,
whole
blood,
plaque,
zinc
phosphate
cement
and
polycarboxylate cement from instruments, metal surfaces and
dentures.
Ultrasonic cleaning minimizes the handling of contaminated
instruments and reduces the chance of injuries from sharp,
contaminated instruments.
Instruments are loaded into a metal basket which is then placed into
the ultrasonic bath. The unit is activated for the time recommended by
manufacturer (≈ 6 minutes). Instruments which are contained in the
cassettes are cleaned for 12 minutes.
After the cleaning cycle is complete, the basket is taken to the sink
and the instruments are carefully and thoroughly rinsed under tap
water. The instruments are checked for residual debris which may be
safely removed manually.
Instruments are taken to the packaging area, where they are unloaded
from the baskets onto a thick disposable paper towel. The instruments
are thoroughly dried using strong paper towels. Drying is important!
Small rotary and endodontic instruments should be held in beakers of
ultrasonic cleaning solution which are suspended in the cleaning bath.
4. Dishwasher instrument cleaners
These machines are popular in Germany and Scandinavia. Dishwasher
cleaners are capable of holding either instrument sets in cassettes or
single instruments in containers.
These machines first pre-rinse the instruments at a low temperature, to
prevent the coagulation of proteins. Washing then takes place at 900C
for 10 minutes using recommended disinfectant/detergent and
instruments are dry after the cycle is completed.
STERILISATION
Instruments must be clean and dry before sterilization.
There are four types of sterilizers used routinely in dentistry:
- The steam autoclave
- The chemical vapor pressure sterilizer
- The dry heat oven
- The glass bead/salt sterilizer
The steam autoclave
A steam autoclave sterilizes by the use of steam under pressure.
Non-vacuum autoclaves are used in dental practice → the use of incoming
steam pressure to evacuate air. Many good quality, modern non-vacuum
autoclaves are very efficient in air removal if the autoclave is not
overloaded.
DENTAL AUTOCLAVES
A prescribed temperature, pressure and time are necessary to destroy
bacterial spores. The temperature-time combinations for autoclaves are
shown in the following tables:
In UK:
Temperature
Pressure
Minimum holding
(0C)
(kPa)
time
134 – 138
207
3 min
121 – 124
103
15 min
In USA:
Temperature
Pressure
Minimum
(0C)
(kPa)
holding time
132
270
3 min
121
103.5
15 min
Lightly wrapped
132
270
8 min
items
121
103.5
20 min
Heavily wrapped
132
270
10 min
items
121
103.5
20 min
Unwrapped items
Items suitable for autoclaving
The steam autoclave is suitable for the sterilization of:
 High-quality stainless steel instruments
 Hand pieces which can be autoclaved
 Cloth goods
 Glass slabs, dishes, stones. Stones should be dry, wet stones
may rupture under steam sterilization.
 Large plastic suction tips
 Packaged and wrapped instruments
 Heat resistant, plastic instruments.
Limitations for use
Carbon-steel instruments will rust after a limited number of
autoclave cycles → prior to autoclaving, instruments susceptible to
rust should be treated with a non-toxic, non-silicone, oil emulsion.
This tends to reduce rust and lubricates instruments without
leaving a noticeable oily or sticky film. These emulsions do not
impair sterilization. Sodium nitrite 2% is used to reduce rusting
and corrosion.
Needles should NOT be autoclaved, nor oil, wax and dry powder!
Practical aspects
Steam autoclave sterilizers are required to:
- Have an instruction manual;
- Have been evaluated for the ability to sterilize, i.e. destroy
bacterial spores;
- Have a drying cycle → dry instruments in trays and packages may
be stored safely for long periods without corrosion.
- The chamber size of the autoclave to be large enough to allow the
sterilization of standard trays.
Directions for use
- Do not overload the trays
- Load the trays carefully to allow free circulation of steam round
the instruments
- Use trays with perforated basis to allow air displacement and free
steam circulation
- Separate wrapped items, do not stack together
- Do not overload the autoclave with trays and pouches
- Sterilization failure in a steam autoclave can usually be avoided if
proper cleansing, packaging and loading precautions are taken
- Remove all particles of amalgam, cements, which adhere to
instruments, after ultrasonic cleaning. This will avoid clogging up
the chamber and prevent impaired operation of the air-valve.
- Rinse disinfectants and cleansers from instruments to avoid the
failure of air-valve and the likelihood of instrument corrosion.
- Sometimes distilled \ deionized water is recommended. In “hard
water” areas, tap water will form scales deposits in the water
chamber.
Autoclave maintenance
- Clean the autoclave chamber weekly with a weak detergent and
rinse well.
- Clean the gasket weekly, applying a recommended silicone
compound to it.
- Check the level of the reservoir twice daily. Do not overfill the
reservoir or fill during the cycle, since this may cause flooding at
the end of the cycle.
- Empty the water reservoir weekly and remove particles on the
chamber floor using high-velocity aspiration.
CHEMICAL VAPOUR STERILISERS
Chemical vapor sterilizers operate by heating a deodorized alcohol,
formaldehyde and ethyl methyl ketone solutions to 132 0C at 138-276 kPa
for 20 minutes in a closed chamber.
Three main advantages:
- A fairly short cycle → 30 minutes. Short “flash cycles” of 7
minutes a re now available on new models when sterilizing
unwrapped items.
- Instruments do not dull, rust or corrode → a longer useful life.
- Instruments are dried at the end of the cycle.
Disadvantages:
- A good ventilation is required to eliminate residual formaldehyde
from the chamber
- An extra cost of purchasing the sterilizing solutions.
A CHEMICLAVE MODEL
Instrument suitable for sterilization with chemical vapor
The chemical vapor sterilizer is suitable for:
- All dental hand instruments
- Hand pieces
- Carbon-steel and diamond burs
- Endodontic instruments
- Orthodontic wires and bands
- Orthodontic pliers, where rusting of the joints may cause problems.
DRY-HEAT STERILISERS
Dry heat is an effective means of sterilization, when it is used
properly.
Unwrapped, moderate loads of instruments placed in an oven can be
sterilized at 160-1700C in 1 hour. Wrapped loads require a longer time of
exposure.
The standard dry-heat oven which is manufactured for dental use (see
picture below) is NOT recommended for a busy dental practice for the
following reasons:
- Low temperature pockets in the chamber during sterilization cycle
may prevent sterilization of the load.
- The complete cycle is very long (≈ 90 minutes) → unsuitable for a
busy dental practice.
- Most ovens do not have a door-locking system, which allows
interruption of the sterilization cycle. If cycle is interrupted, loads are not
sterile.
- Turbine hand pieces are damaged if heated to a temperature of
1600C
THE DRY-HEAT OVEN
Rapid dry-heat sterilization
The Cox rapid dry-heat oven operates by circulating air at a steady
temperature of 1900C. At this temperature, most metal instruments are not
damaged.
Sterilization time:
6 minutes for unwrapped instruments
12 minutes for wrapped instruments
The load size of this sterilizer is comparatively small.
HOT BEAD / SALT STERILIZERS
Glass bead/salt sterilizers are useful for sterilizing small instruments:
- Endodontic files
- Rotary instruments
The sterilizer is a metal cup in which the medium is maintained at 218
- 246 0C.
It is useful for sterilizing endodontic files during endodontic treatment
of multirooted teeth, to prevent transfer of infection between canals.
A minimum of 15 seconds immersion is recommended for effective
sterilization. The instruments to be sterilized should be thoroughly clean.
Hot salt sterilizers are preferably to glass bead sterilizers for
endodontic therapy. Residual salt or glass beads on the surface of the sterile
endodontic instruments can cause blockage of the root canals. If salt
granules are the cause of the blockage, they can be dissolved with irrigation.
A suitable thermometer should be routinely used to ensure that the
temperature reaches 2180C, as the temperature of the medium at the centre
of the sterilizer may be lower than that at the sides of the container.
MONITORING STERILIZATION
A high sterilization failure in dental sterilizers (≈ 51%)
Positive results may be caused when:
- Packs are improperly wrapped
- The sterilizer is improperly loaded
- The sterilizer does not work properly
- The process time is too short
1. Tests using chemical indicators
Test strips and tubes are available for autoclaves, chemiclaves and dry-heat
ovens → very useful for early detection of problems.
Chemical indicators show that sterilizing conditions have been reached, not
that sterilization has occurred.
Color-change strips or tapes → indicate temperature change
Superiors color-change indicators → sensitive to both temperature and time
Indicator strip before use
After successful sterilization → color change
2. Biological monitors
Self-contained biological indicators: bacterial spores on a paper
carrier packaged within a small, thermoplastic culture tube. Inside the
thermoplastic culture tube is a sealed glass ampoule of specially formulated
soybean casein digest culture medium containing a color indicator that turns
a dramatic yellow when spores grow.
Advantages: Easy-to-use; produces visual results within 48 hours
without laboratory transfers.
Download