TEST REPORT IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic disturbances Requirements and tests Report Number. .............................. : TGM-VA EE 36724b-1 SFT Date of issue ................................... : 13.08.2018 Total number of pages ................... 69 Name of Testing Laboratory preparing the Report ..................... : Staatliche Versuchsanstalt ........................... : Address ........................................... : TGM Elektrotechnik und Elektronik Siemens AG Austria | SIMEA Siemensstrasse 90, 1210 Vienna, Austria Test specification: Standard.......................................... : IEC 60601-1-2:2014 Test procedure ............................... : Type test only Non-standard test method ............ : N/A Test Report Form No. .................... : IEC60601_1_2E_EMC Test Report Form(s) Originator .... : UL(US) Master TRF...................................... : 2017-03 NO CB TEST REPORT Copyright © 2017 IEC System of Conformity Assessment Schemes for Electrotechnical EQUIPMENT and Components (IECEE System). All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resul ting from the reader's interpretation of the reproduced material due to its placement and context. If this Test Report Form is used by non-IECEE members, the IECEE/IEC logo and the reference to the CB Scheme procedure shall be removed. This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory and appended to a CB Test Certificate issued by an NCB in accordance with IECEE 02. General disclaimer: The test results presented in this report relate only to the object tested and documents reviewed. This report shall not be reproduced, except in full, without the written approval of the Issuing CB Testing Laboratory. The authenticity of this Test Report and its contents can be verified by contacting the NCB, responsible for this Test Report. Page 2 of 69 TGM-VA EE 36724b-1 SFT Test item description ....................... : switch mode power supply Trade Mark ........................................ : Siemens AG Austria Manufacturer .................................... : Leco GmbH, Franz Jonas Platz 10-12/2/10, 1210 Vienna, Austria Model/Type reference ...................... : 6EP4436-0SB00-0AY0 Ratings .............................................. : Input: 3AC 400-500V | 1,2A | 50-60Hz ; Output: DC 24V | 20A; Responsible Testing Laboratory (as applicable), testing procedure and testing location(s): Testing Laboratory: Staatliche Versuchsanstalt TGM Elektrotechnik und Elektronik Testing location/ address............................. : 1200 Wien, Wexstrasse 19-23, Austria Tested by (name, function, signature) ........ : CBTL W. Baumgartner Approved by (name, function, signature) ... : CBTL T. Thun Testing procedure: CTF Stage 1: Testing location/ address............................. : Tested by (name, function, signature) ........ : Approved by (name, function, signature) ... : Testing procedure: CTF Stage 2: Testing location/ address............................. : Tested by (name, function, signature) ........ : Witnessed by (name, function, signature) . : Approved by (name, function, signature) ... : Testing procedure: CTF Stage 3: Testing procedure: CTF Stage 4: Testing location/ address............................. : Tested by (name, function, signature) ........ : Witnessed by (name, function, signature) . : Approved by (name, function, signature) ... : Supervised by (name, function, signature) : TRF No. IEC60601_1_2E_EMC Page 3 of 69 TGM-VA EE 36724b-1 SFT List of Attachments (including a total number of pages in each attachment): Summary of testing: Tests performed (name of test and subclause): Conducted EMISSIONS (7.3) Radiated EMISSIONS (7.3) Disturbance Power EMISSIONS (7.3) Harmonic Currents (7.2.1) Voltage Fluctuations and Flicker (7.2.3) Testing location: Staatliche Versuchsanstalt TGM Elektrotechnik und Elektronik Wexstrasse 19-23 1200 Vienna, Austria Electrostatic Discharges (8.9) RF Electromagnetic Fields (8.9) Electrical Fast Transients (8.9) Surge (8.9) Conducted Disturbances Induced by RF Fields (8.9) Voltage Dips and Interruptions (8.9) Note: delete tests not conducted in the list above. Summary of compliance with National Differences: EN 60601-1-2:2014 The product fulfils the requirements assumed the fulfilling of the requirements of all applicable standards of all components with only manufactured declaration of conformity according to the mentioned standards / European directive. Device modifications necessary for compliance: No modification necessary (Describe ME EQUIPMENT or ME SYSTEM modifications needed in order to pass any of the EMISSIONS or tests IMMUNITY A statement that manufacturer will incorporated all changes into production units. Manufacturer provided declaration statement. Manufacturer did not provide declaration statement. TRF No. IEC60601_1_2E_EMC