CALIBRATION REPORT
CBScan PA0304
Crest Test Systems(P)Ltd.
30/06/2021
CAL/PGL001
PRODUCT
MANUFACTURER
CALIBRATED ON
REPORT NO.
TESTED / C.4L. BY
SR.NO.
CALIBRATION DUE ON
CUSTOMER
Prashant
5503.00.L-21/001
29/06/2022
OLG,Chennai
1.0 Cbscan PA0304
1.1 OBSERVATIONS :
1. Laboratory Ambients Temperature : 28°C
Sr.
No.
Parameter
Value on Calibration
Standard.
01
TIME
INTERVAL
100 ms.
02.
VELOCITY
1.0 Mls
03.
DC CURRENT
1.0 A.
Value Measured on UUC
Deviation
Allowed
R1 - 100.0 ms
Y1 -100.1 ms
BI - 100.0 ms
CHANNEL TR - 1.01 M/s.
CHANNEL TY - 1.00 M/s.
CHANNEL TB - 1.00 M/s.
CHANNEL CC - 1.00 Amp
+ 1 ms.
+ 1 ms.
+ 1 ms.
+ 0.1 M/s.
+ 0.1 M/s.
+ 0.1 M/s.
+ 10 mA.
The product under test complies with calibration specification indicated above and is verified
to be within calibration
1.2 EQUIPMENT USED FOR CALIBRATION :
1. Crest Test Systems(P)Ltd. make SCALE TCV calibrator (Sr. No. 5509.00.A002)
* Traceability report of above instrument is attached along with.
Checked By
Issued By
Calibration
Status
OK
OK
OK
OK
OK
OK
OK