FITTING TECHNIQUE Diameter Selecting Total Diameter • Individual practitioner preference • Independent variables to consider: ⎼ How much clearance is required- larger diameter lenses offer larger sagittal height ⎼ Greater bearing area on the sclera/conj therefore less pressure on conj – better comfort ⎼ Smaller lenses “sink into the cornea more – also less movement of the lens ⎼ Larger lenses tend to decenter more temporally (due to flatter nasal sclera) ⎼ Smaller lenses are easier to handle SA LENSES - GUIDE LENS HVID(mm) CLLSA: 2-Spec/SSF REC DIAMETER(mm) 15.6 (SSF) 16.7 (SW) 17.0 21.0 CLLSA – DF CLLSA: Rose –K2 XL Danker Lenses: Zenlens INNOCON: Eye Space Scleral From 15.8 11.8 (1.3-1.5>) 13.6-15.6 Average 14.60 <11.7 16.00 >11.8 17.00 < 11.5 16.50 >= 11.5 17.50 FITTING TECHNIQUE Optical Zone/Clearance Zone Diameter • Fixed optic zone diameters –may not be possible to adjust this parameter in some lens designs. • Consider: Pupil diameter taking into account the depth of the AC and possible lens decentration • The clearance zone area (optical zone + limbal zone) should be chosen approx. 0.2mm larger than the corneal diameter • If the optical zone diameter cannot be adjusted – change the overall lens design/larger diameter lens FITTING TECHNIQUE: CLEARANCE How to determine corneal clearance • no rules - typically a minimum of 100-150 (vs 300) microns is desired • Some cases - lenses can be fitted with a minimum of 20-50 microns of clearance • In larger lenses 200-300 microns is considered sufficient however this can be as much as 500-600 microns • Keratoconics need higher sag than a post-graft or a post-refractive surgery. • ocular surface disease - larger sag FITTING TECHNIQUE: CLEARANCE • Start with a low/high sag height – adjust until you have sufficient clearance – approx. 150-300 microns • Fill lens cup with preservative-free saline & plenty fluorescein • No air bubbles • An ideal optical zone fit is an equal green fluorescein pattern with no areas of bearing (no black patches) • Fluorescein becomes visible to the human eye at a minimum tear lens thickness of 20 microns. • “Black” indicates insufficient clearance/bearing FITTING TECHNIQUE: CLEARANCE FITTING TECHNIQUE: CLEARANCE • If central corneal “bearing” is visible this means the sag of the lens may be too low. • If bubbles appear under the lens (not from incorrect insertion) – maybe excessive corneal clearance – lens edge lifting • Bubbles also from nonsymmetrical shape of the ant. ocular surface. • Small bubbles that move may be acceptable provided they do not cross the pupil • large stationery bubbles are not acceptable. The ideal fit – no bubbles. FITTING TECHNIQUE: CLEARANCE • Excessive clearance, of >500 microns -decrease in VA & increased visual disturbances. • keratoconus - higher sagittal height depending on cone height • “feather touch” with marginal bearing on the cornea - Fitting goal is minimal sagittal height, vault the cornea with marginal or no corneal bearing. • Minimal bearing appears to be better tolerated than with corneal RGPs – minimal lens movement less irritation to corneal apex. • The lenses need a minimum settling time of approx. 3040min before evaluating FITTING TECHNIQUE: CLEARANCE • The lenses may “sink” up to 100 microns - therefore allow adequate settling time before assessing • Fenestrated lenses may sink more than non-fenestrated lenses • Use the optic section at 45º angle, on your slit lamp to evaluate the post-lens corneal tear film thickness – white light FITTING TECHNIQUE: CLEARANCE Peripheral/Limbal Clearance • BC slightly flatter than the flattest K value helps to create clearance in the peripheral/limbal zone • flattening the BC will reduce sagittal height as well - need to compensate by increasing the sag height slightly. • Most manufacturers factor this in by default • Sag height is also dependent on the lens diameter. Increasing the diameter for a particular BC will result in an increase in the sag height. - manufacturers compensate for this FITTING TECHNIQUE: CLEARANCE • No bearing on limbus - limbal stem cells • Limbal stem cells – regeneration of corneal epithelium • Clearance of approx. 100 microns – NO staining over limbus • corneo-scleral lenses do not always clear limbus adequately, however some lens designs come with different transitions zone profiles. • Persistent bubble in limbal zone - eliminate by flattening the BC or fitting a lower profile limbal zone curve. An OCT can be extremely useful in determining a more precise clearance and aiding in fitting. A B SCLERAL LANDING ZONE( SLZ) • The scleral landing zone is closely related to the amount of clearance that is created beneath the lens. • If the SLZ is steep it will lift the lens creating more clearance. • If there is marked central corneal touch the landing zone will be elevated off the ocular surface – edge lift • The goal is to create alignment with the sclera/corneo-scleral zone SCLERAL LANDING ZONE • The alignment can be assessed using a slit lamp or using an OCT • Once the trial lens has settled assess the fit based on how the landing zone bears on the sclera/conj. • A ring of bearing on the inner aspect of the landing zone indicates that the SLZ is too flat. The presence of air bubbles or frothing in the periphery also indicates this- heel blanching • Bearing on the outer aspect of the SLZ indicates a steep fit and will increase the overall vault of the lens – toe blanching • Areas of “whitening” indicate blanching of conjunctival vessels due to compression from the landing curve. Blanching of conjunctival vessels on the outer aspect of the landing zone i.e. toe blanching Anterior segment 5-line raster images show a scleral lens edge that lands too flat and lifts away from the conjunctiva (A); good, stable landing aligned with conjunctiva (B); and impingement into the conjunctiva with a steep landing zone edge (C). • Assess for blanching in all directions of gaze. • Conj compression does not necessarily result in conjunctival staining on lens removal but can cause rebound hyperaemia. • If the lens is ‘pinching’ on the conjunctival tissue this impingement and will stain on lens removal. Long term impingement can result in conjunctival hypertrophy (thickening). • The larger the lens diameter the more lens weight is distributed over a larger area of the sclera. This results in the scleral lens vaulting the cornea more. • High vault may allow for limited movement of the lens. LENS EDGE • Too much edge lift may cause discomfort and increased lens awareness • Low edge profiles can cause impingement on the conjunctiva and obstruct blood flow • Observe the edge using white light. Check whether it sinks into the conjunctiva or whether it lifts. • Edge lift will be observed as a dark ring around the edge of the lens • Impinging will be observed as conjunctival staining on lens removal. • Most lens manufacturers allow for a degree of variation in the landing zone LENS EDGE Non-rotationally Symmetrical Lens Design Fitting toric scleral lenses • As mentioned previously the anterior ocular surface of the eye is non-rotationally symmetric. • This implies that one or two aspects of the sclera/conjunctiva are compressed more when a scleral lens is placed on the eye. • Different methods of alleviating this compression/impingement have been used/suggested e.g. truncating the lens • Back toric or quadrant specific lens designs are an alternative to alleviating this problem, improving the lens fit and comfort. LENS EDGE • The toric or quadrant specific curve is applied only to the SLZ • Makes the lens more susceptible to flexure and therefore should be used if the anterior ocular surface shows significant irregularity. • Toric scleral lenses allow for more equal distribution of pressure over the corneal surface, Visser et al (2006) • The difference in the sag height between the two curves can vary between 100-1000 microns and is usually determined by the lab/manufacturer • Evaluation of a toric scleral lens is the same as for a spherical lens. • Once an optimal fit is obtained perform an over-refraction. Fitting quadrant specific lenses • Usually an empirical approach is used to fit quadrant specific designs • Use a standard fitting set, determine the area of lift off at the edge using an optic section and a reference such as the central corneal thickness. • Theoretically the lens should find its own orientation on the cornea however since these lenses don’t move much lens markings help to insert the lens correctly • If more than one quadrant is altered then the location of the specific quadrants need to be indicated. • If needed/desired front optics can be applied, as with toric RGPs Fitting front toric scleral lenses • If the over refraction requires a cyl while no toric back surface is needed, the lens is considered a true front toric lens and needs to be stabilized some how. Prism ballast /double slab-off ballasting is used to stabilize the lens. • As with standard lenses use the LARS rule to assess the rotation of the lens. • Scleral lenses typically don’t move however larger lenses may move slightly more than smaller lenses Movement• The lens should be mobile when pushed up slightly • Changing the lens edge does not seem to have too much impact on lens movement. • If the lens has too little apical clearance it may vault across the corneal apex or along the flat meridian on a toric ocular surface. • Optimize the lens fit first before performing an overrefraction. • If the over rx exceeds 4.00 D convert for the vertex distance correction • Trial frame over-refraction