1 M-18D DILUENT 【PRODUCT NAME】 M-18D DILUENT 【MODEL】 M-18D 【PACKAGE SPECIFICATIONS】 5.5L×2、20L 【INTENDED USE】 For use on MINDRAY BC-1800/BC-2600/BC-2800 hematology analyzer. For use as an azide-free, filtered isotonic solution for counting and sizing blood cells. Only use one time. Don 't use reagent out off date. 【ACTIVE INGREDIENTS】 Sodium Chloride………………………………..……....…3.0-5.5 g/L Sodium Sulfate Anhydrous………………………...…...7.5-11.5 g/L Buffering Agents………………….………………..….…..1.0-3.0 g/L Anti-fungal and Anti-bacterial Agents………………..….0.8-2.5 g/L 【STORAGE CONDITION AND SHELF LIFE】 The product must be stored in the temperature of 2℃~30℃(35℉~86℉), The shelf life is 24 Months. The product must be used in the temperature of 15℃~30℃(59℉~86℉), and it stays valid for 60 days after being opened. 【APPLICABLE ANALYZER MODELS】 The product applies to Mindray BC-1800/BC-2600/BC-2800 Hematology analyzer manufactured by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.. 【INSTRUCTIONS FOR USE】 1 Keep Hematology Analyzer Reagent between 15℃-30℃ during use. 2 Refer to Mindray Operation Manual to obtain instrument-specific instructions for appropriate use of Hematology Analyzer Reagent. 【WARNINGS AND PRECAUTIONS】 If the reagent accidentally gets into your mouth, seek medical assistance immediately. If the reagent spills on your skin, wash it off with plenty of water immediately. If the reagent spills into your eyes, wash it off with plenty of water immediately. 【NOTICE】 Do not use expired products. If the product has been frozen(either partially or completely), warm it up to room temperature and then mix it by gentle inversion. Verify background results before analyzing patient samples.If background results of the reagent are abnormal due to transportation, keep the reagent still in room temperature for 24 hours before use. If reagents are contaminated or become abnormal due to other reasons, stop using and then replace them at once. Be sure to dispose of reagents, waste and contaminated packing materials according to government regulations. The following conditions may produce interference to test results: expired reagent; reagent contaminated by dusts; unclean sa mple dilution; reagent mixed with those of other manufacturers; residue reagent mixed with newly opened reagent; reagent not used as required. 【CUSTOMER SERVICE DEPARTMENT】 Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray building, Keji 12th Road South, Hi-tech Industrial Park,Nanshan, Shenzhen, 518057 P.R.China Website: www.mindray.com E-mail Address: service@mindray.com Tel: +86 755 81888998 Fax: +86 755 26582680 © 2014 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.For this Operator’s Manual, the issue date is 2014-02. In vitro diagnostic medical Batch code Use -by date Temperature limit Manufacture device P/N:046-006108-00-1.0 M-23D DILUENT 02 【PRODUCT NAME】 M-23D DILUENT 【MODEL】 M-23D 【PACKAGE SPECIFICATIONS】 10L、20L 【INTENDED USE】 For use on Mindray BC-2300/BC-2100 hematology analyzer. For use as an azide-free, filtered isotonic solution for counting and sizing blood cells. Only use one time. Don’t use reagent out off date. 【ACTIVE INGREDIENTS】 Sodium Chloride………………………………..……....…3.0-5.5 g/L Sodium Sulfate Anhydrous………………………...…...7.5-11.5 g/L Buffering Agents………………….………………..….…..1.0-3.0 g/L Anti-fungal and Anti-bacterial Agents………………..….0.8-2.5 g/L 【STORAGE CONDITION AND SHELF LIFE】 The product must be stored in the temperature of 2℃~30℃(35℉~86℉), The shelf life is 24 Months. The product must be used in the temperature of 15℃~30℃(59℉~86℉), and it stays valid for 60 days after being opened. 【APPLICABLE ANALYZER MODELS】 The product applies to Mindray BC-2300/BC-2100 Hematology analyzer manufactured by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.. 【INSTRUCTIONS FOR USE】 1 Keep Hematology Analyzer Reagent between 15℃-30℃ during use. 2 Refer to Mindray Operation Manual to obtain instrument-specific instructions for appropriate use of Hematology Analyzer Reagent. 【WARNINGS AND PRECAUTIONS】 If the reagent accidentally gets into your mouth, seek medical assistance immediately. If the reagent spills on your skin, wash it off with plenty of water immediately. If the reagent spills into your eyes, wash it off with plenty of water immediately. 【NOTICE】 Do not use expired products. If the product has been frozen(either partially or completely), warm it up to room temperature and then mix it by gentle inversion. Verify background results before analyzing patient samples.If background results of the reagent are abnormal due to transportation, keep the reagent still in room temperature for 24 hours before use. If reagents are contaminated or become abnormal due to other reasons, stop using and then replace them at once. Be sure to dispose of reagents, waste and contaminated packing materials according to government regulations. The following conditions may produce interference to test results: expired reagent; reagent contaminated by dusts; unclean sample dilution; reagent mixed with those of other manufacturers; residue reagent mixed with newly opened reagent; reagent not used as required. 【CUSTOMER SERVICE DEPARTMENT】 Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray building, Keji 12th Road South, Hi-tech Industrial Park,Nanshan, Shenzhen, 518057 P.R.China Website: www.mindray.com E-mail Address: service@mindray.com Tel: +86 755 81888998 Fax: +86 755 26582680 © 2010 – 2016 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.For this Operator’s Manual, the issue date is 2016-08. In vitro diagnostic medical Batch code Use-by date Temperature limit Manufacture device M-30D DILUENT 03 Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Mindray are the trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational or editorial purposes. They are the property of their respective owners. 1. Product Name M-30D DILUENT. 2. Intended Use For use on Mindray BC-20/BC-21/BC-10/BC11/BC-3000Plus/BC-3000CT/BC-3200/BC3300/BC-3600/ BC-20s/BC-21s/BC-30s/ BC-31s/BC-30/BC-31/BC-30e/BC-10e hematology analyzer. For use as an azide-free, filtered isotonic solution for counting and sizing blood cells. Only use one time. Don’t use reagent out off date. 3. Active Ingredients Sodium Chloride…………………………………3.0-5.5g/L Sodium Sulfate Anhydrous……………………7.5-11.5g/L Buffering Agents…………………………………1.0-3.0g/L Anti-fungal and Anti-bacterial Agents…… …...0.8-2.5g/L 4. Warnings and Precautions Do not inhale and/or ingest. In case of inhalation or ingestion, seek medical advice immediately. Avoid skin contact. In case of skin contact, flush the affected area with plenty of water immediately. Avoid eye contact. In case of eye contact, flush the affected area with plenty of water and seek medical advice immediately. 5. Storage and Usage The product must be stored in the temperature of 2℃ to 30℃(35℉ to 86℉ ), and the relative humidity must not exceed 90%. The storage room must be well ventilated and without corrosive gas. The shelf life is 2 years. The product must be used in the temperature of 15℃ to 30℃(59℉ to 86℉ ), and it stays valid for 60 days after being opened. Do not use expired products. If product has been frozen(either partially or completely), warm it up to room temperature and then mix it by gentle inversion. You can only proceed to test samples after you have confirmed that the measured blank value is within the required range. Dispose of waste product, unused product and contaminated packaging in accordance with governmental requirements. The following conditions may produce interference to test results: expired reagent, reagent contaminated by dusts, unclean sample dilution, reagent mixed with those of other manufacturers Warning If the reagent accidentally gets into your mouth, seek medical assistance immediately; if the reagent spills on your skin, wash it off with plenty of water immediately; if the reagent spills into your eyes, wash it off with plenty of water and seek medical assistance immediately Note: This product must only be used by Mindray authorized personnel or skilled/trained clinical professionals. GRAPHICAL SYMBOLS M-50 SERIES HEMATOLOGY REAGENTS PRODUCT NAME M-50D DILUENT 4 M-50 LEO (I) LYSE M-50 LEO (II) LYSE M-50 LH LYSE M-50 LBA LYSE M-50 CLEANSER PROBE CLEANSER 16 22 SUMMARY AND PRINCIPLE HEMATOLOGY ANALYZER REAGENT is a series of reagents which are formulated specially for Mindray hematology analyzers. The reagents are M-50D DILUENT, M-50 LEO (I) LYSE , M-50 LEO (II) LYSE ,M-50 LH LYSE,M-50 LBA LYSE,M-50 CLEANER, PROBE CLEANSER. M-50D DILUENT is an azide-free, filtered isotonic solution used for counting and sizing blood cells. M-50LH Lyse is used to break down red blood cell membrane and converts hemoglobin to a hemoglobin complex to determine the HGB. M-50LBA Lyse is used to break down red blood cell membrance and 2-differentiates WBCs to basophils and other WBCs. Use the M-50LBA Lyse as instructed by the operator’s manual of the analyzer. M-50LEO(I) Lyse breaks down the red blood cell and cooperates with M-50LEO(II) Lyse to 4-differentiate WBCs. M-50LEO(II) Lyse cooperates with M-50LEO(I) Lyse to 4-differentiate WBCs. M-50 Cleanser is used for shutdown and regular cleaning. PROBE CLEANSER is used to clean the analyzer regularly. INTENDED USE For In Vitro Diagnostic use only one time. Don't use reagent out off date. INSTRUCTIONS FOR USE Refer to Mindray Operator’s manual to obtain instrument-specific instructions for appropriate use of HEMATOLOGY ANALYZER REAGENT. Keep Hematology Analyzer Reagent between 15℃-30℃during use. STORAGE AND STABILITY M-50D DILUENT, M-50 LEO (I) LYSE , M-50 LEO (II) LYSE ,M-50 LH LYSE,M-50 LBA LYSE have a shelf life of 24 months when stored at 2-30℃. Probe cleanser has 12months shelf life under 2-30℃. The open vial life for all the reagents is 60days when the operating environment temperature is 15℃-30℃. PROBE CLEANSER needs to be stored in a dark place. If frozen, the product needs to be thaw completely before mixing. Mix the product by gentle inversion prior to initial placement on instrument. Verify acceptable background counts before performing patient sample analysis. WARNINGS AND PRECAUTIONS Do not ingest, keep away from foods. Consult a physician if you accidentally ingest it. Avoid contact with skin and eyes. In case of skin contact, immediately flush with water. In case of contact with eyes, rinse immediately with plenty of water and seek medical attention immediately. M-50P PROBE CLEANSER includes irritating substances. Please keep the container closed when not in use. ACTIVE INGREDIENTS M-50LEO(I) Lyse Surfactant <35g/L M-50LEO(II) Lyse Surfactant <40g/L M-50LH lyse Quarternary Ammonium Salts <50g/L Nonionic Surfactant Isopropanol Ethanol <10g/L 0.1-1.5mL/L <1.5mL/L M-50LBA lyse Surfactant <50g/L M-50 Cleanser Surfactant Buffering agents Detergent enzyme 0.5-2.0g/L 0.5-3.0g/L 0.05-0.5g/L Anti-fungal and anti-bacterial agents 0.01-1.0g/L PROBE CLEANSER Surfactant Sodium hypochlorous Sodium hydroxide <2.0g/L <100g/L <50g/L MANUFACTURER Shenzhen Mindray Bio-Medical Electronics Co.,Ltd. Mindray Building , Keji 12th Road South , Hi-tech Industrial Park , Nanshan , Shenzhen P.R.China,518057 AUTHORIZED REPRESENTATIVE EUROPEAN COMMUNITY Shanghai International Holding Crop. GmbH(Europe) Eiffestrasse 80 D-20537 Hamburg Germany 5 M-52D DILUENT 【PRODUCT】 This product applies to BC-5100/BC-5300/BC-5180/ BC5380/BC-5310/BC-5390/BC-5390 CRP/BC-5180 CRP M-52D DILUENT Auto Hematology Analyzer produced by Shenzhen Mindray 【PACKAGE SIZE】 Bio-Medical Electronics Co., Ltd. 20L 【INSTRUCTIONS FOR USE】 【INTENDED USE】 M-52D diluent provides a stable environment for counting and sizing blood cells. Only use one time. Don't use reagent out off date. Unpack the reagent container. Insert the pickup pipe from the analyzer's cap assembly into the container, making sure the container's cap and the connected assembly 【PRINCIPLE】 M-52D diluent provides stable solution environment for connector are in the same color. Turn and secure the cap blood cells, dilutes blood sample, generates sheath fluid assembly tightly, and replace the reagent in accordance and provides conductance in blood cell analysis process. with the instructions in the Operator's Manual of the analyzer. You can only proceed to test samples after you 【ACTIVE INGREDIENTS】 Sodium Chloride……………………………………3.0-5.5g/L Sodium Sulfate Anhydrous………………………7.5-11.5g/L Buffering Agents……………………………………1.0-3.0g/L Anti-fungal and Anti-bacterial Agents………...…0.8-2.5g/L 【PRODUCT SPECIFICATIONS】 1. Appearance: Transparent liquid without deposits, required range. For detailed usage instruction, please refer to the Operator's Manual of the analyzer. 【WARNINGS】 Do not inhale and/or ingest. In case of inhalation or ingestion, seek medical advice immediately. Avoid skin contact. In case of skin contact, flush the affected area suspended grains or flocks. 2. have confirmed that the measured blank value is within the Blank count results: The blank count results tested on the matched Auto Hematology Analyzer shall meet the with plenty of water immediately. Avoid eye contact. In case of eye contact, flush the affected area with plenty of water and seek medical advice immediately. requirements of Table 1. Table 1 Blank Count Requirements Parameter Blank Count Requirements WBC ≤0.3×109/L RBC ≤0.03×1012/L PLT ≤10×109/L HGB ≤1 g/L 【NOTICE】 1. Do not use expired products. 2. If product has been frozen(either partially or completely),warm it up to room temperature and then mix it by gentle inversion. 3. Verify back ground results before analyzing patient samples. 4. Dispose of waste product, unused product and 【STORAGE AND EXPIRATION DATE】 contaminated This product will be stable for 2 years when stored at 2℃to 30℃(35℉to 86℉). The product should be used at 15℃to 30℃(59℉to 86℉) once opened, and it is stable for 60 days. 【APPLICABLE INSTRUMENT】 packaging governmental requirements. in accordance with English 1/2 00(1.0) 【GRAPHICAL SYMBOLS】 Batch code Use-by date 【COMPANY CONTACT】 Temperature limit In vitro diagnostic medical device Manufacture Authorized representative in the European Community Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen 518057, P.R.China Website: www.mindray.com E-mail Address: service@mindray.com Tel: +86 755 81888998 Fax: +86 755 26582680 EC-Representative: Shanghai International Holding Corp. GmbH(Europe) Address: Eiffestraβe 80, Hamburg 20537, Germany Tel: 0049-40-2513175 Fax: 0049-40-255726 © 2014 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved. 【APPROVAL DATE OF THE OPERATOR’S MANUAL】 2014-07 English P/N: 046-006524- 2/2 P/N: 046-006524-00(1.0) 6 M-53D DILUENT 【PRODUCT】 This M-53D DILUENT product applies to BC-5100/BC-5300/BC-5180/ BC-5380/BC-5310/BC-5390/BC-5390 CRP/BC-5180 CRP 【PACKAGE SIZE】 Auto Hematology Analyzer produced by Shenzhen 20L Mindray Bio-Medical Electronics Co., Ltd. 【INTENDED USE】 【INSTRUCTIONS FOR USE】 M-53D diluent provides a stable environment for counting Unpack the reagent container. Insert the pickup pipe from and sizing blood cells. Only use one time. Don't use reagent out off date. 【PRINCIPLE】 the analyzer's cap assembly into the container, making M-53D diluent provides stable solution environment for connector are in the same color. Turn and secure the cap blood cells, dilutes blood sample, generates sheath fluid assembly tightly, and replace the reagent in accordance and provides conductance in blood cell analysis process. with the instructions in the Operator's Manual of the 【ACTIVE INGREDIENTS】 sure the container's cap and the connected assembly analyzer. You can only proceed to test samples after you Sodium Chloride……………………………………3.0-5.5g/L have confirmed that the measured blank value is within the Sodium Sulfate Anhydrous………………………7.5-11.5g/L required range. For detailed usage instruction, please refer Buffering Agents……………………………………1.0-3.0g/L to the Operator's Manual of the analyzer. 【WARNINGS】 Anti-fungal and Anti-bacterial Agents………...…0.8-2.5g/L 【PRODUCT SPECIFICATIONS】 Do not inhale and/or ingest. In case of inhalation or 1. Appearance: Transparent liquid without deposits, ingestion, seek medical advice immediately. suspended grains or flocks. Avoid skin contact. In case of skin contact, flush the 2. Blank count results: The blank count results tested on affected area with plenty of water immediately. the matched Auto Hematology Analyzer shall meet the Avoid eye contact. In case of eye contact, flush the requirements of Table 1. affected area with plenty of water and seek medical advice immediately. Table 1 Blank Count Requirements Parameter Blank Count Requirements WBC ≤0.3×109/L RBC ≤0.03×1012/L PLT ≤10×109/L HGB ≤1 g/L 【NOTICE】 1. Do not use expired products. 2. If product has been frozen(either partially or completely),warm it up to room temperature and then mix it by gentle inversion. 3. Verify back ground results before analyzing patient 【STORAGE AND EXPIRATION DATE】 This product will be stable for 2 years when stored at 2℃ samples. to 30℃(35℉ to 86℉). 4. Dispose of waste product, unused product and The product should be used at 15℃ to 30℃(59℉ to 86℉) contaminated packaging in accordance with governmental once opened, and it is stable for 60 days. requirements. 【APPLICABLE INSTRUMENT】 English 1/2 P/N: 046-006524-00(1.0) 【GRAPHICAL SYMBOLS】 Batch code Use-by date Temperature limit In vitro diagnostic medical device Manufacture Authorized representative in the European Community 【COMPANY CONTACT】 Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen 518057, P.R.China Website: www.mindray.com E-mail Address: service@mindray.com Tel: +86 755 81888998 Fax: +86 755 26582680 EC-Representative: Shanghai International Holding Corp. GmbH(Europe) Address: Eiffestraβe 80, Hamburg 20537, Germany Tel: 0049-40-2513175 Fax: 0049-40-255726 © 2014 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved. 【APPROVAL DATE OF THE OPERATOR’S MANUAL】 2014-07 English 2/2 P/N: 046-006524-00(1.0) 7 M-58D Diluent 【PRODUCT】 M-58D DILUENT 【PACKAGE SIZE】 20L 【INTENDED USE】 M-58D Diluent provides a stable environment for counting and sizing blood cells. Only use one time. Don't use reagent out off date. 【ACTIVE INGREDIENTS】 Sodium Chloride……………………………………3.0-5.5g/L Sodium Sulfate Anhydrous………………………7.5-11.5g/L Buffering Agents……………………………………1.0-3.0g/L Anti-fungal and Anti-bacterial Agents………....…0.8-2.5g/L 【PRODUCT SPECIFICATIONS】 1. Appearance: Transparent liquid without deposits, suspended grains or flocks. 2. Blank count results: The blank count results tested on the matched Auto Hematology Analyzer shall meet the requirements of Table 1. Table 1 Blank Count Requirements Parameter Blank Count Requirements WBC ≤0.3×10 /L RBC ≤0.03×10 /L PLT ≤10×10 /L HGB 1g/L 9 12 9 【STORAGE AND EXPIRATION DATE】 The product will be stable for 2 years when stored at 2℃ to 30℃(35℉ to 86℉). The product should be used at 15℃ to 30℃(59℉ to 86℉) once opened, and it is stable for 60 days. 【APPLICABLE INSTRUMENT】 This product applies to BC-5600 and BC-5800 Auto Hematology Analyzer produced by Shenzhen Mindray BioMedical Electronics Co., Ltd. 【INSTRUCTIONS FOR USE】 Unpack the reagent container. Insert the pickup pipe from the analyzer's cap assembly into the container, making sure the container's cap and the connected assembly connector are in the same color. Turn and secure the cap assembly tightly, and replace the reagent in accordance with the instructions in the Operator's Manual of the analyzer. For detailed usage instruction, please refer to the Operator's Manual of the analyzer. 【WARNINGS】 Do not inhale and/or ingest. In case of inhalation or ingestion, seek medical advice immediately. Avoid skin contact. In case of skin contact, flush the affected area with plenty of water immediately. Avoid eye contact. In case of eye contact, flush