Service Manual Eleganza 2 Smart acute bed D9S001GRA-0101 Version: 07 Publication Date: 2021-07 2 D9S001GRA-0101 Table of Content 1 Introduction - General information...................................................................................................... 5 1.1 1.2 Product and technical label description and location...................................................................... 5 Identification of bed sides................................................................................................................ 6 2 Technical data and Symbols................................................................................................................. 7 3 Theory of Operations............................................................................................................................ 8 3.1 3.1.1 3.1.2 3.1.3 3.2 3.3 3.3.1 3.3.2 3.4 3.4.1 3.4.2 3.4.3 3.4.4 3.4.5 3.5 3.5.1 3.6 3.7 3.7.1 3.8 3.8.1 3.8.2 Control box...................................................................................................................................... 8 Power unit....................................................................................................................................... 9 Control unit...................................................................................................................................... 9 Battery pack.................................................................................................................................. 12 Column units................................................................................................................................. 13 Linear actuators............................................................................................................................. 14 Backrest actuator.......................................................................................................................... 14 Thighrest actuator......................................................................................................................... 14 Controllers..................................................................................................................................... 15 Attendant control panel (ACP)....................................................................................................... 15 Handset......................................................................................................................................... 15 Foot controller (height elevation)................................................................................................... 16 Siderail controls (optional)............................................................................................................. 16 Satellite (optional).......................................................................................................................... 17 Night illumination........................................................................................................................... 18 Undercarriage module – Basic...................................................................................................... 18 Brake alarm (optional)................................................................................................................... 19 Castors.......................................................................................................................................... 19 Mechanical 5th castor................................................................................................................... 20 Bed wiring diagram........................................................................................................................ 21 Bed wiring diagram (universal for PB43)....................................................................................... 21 Bed wiring diagram (universal for PB21)....................................................................................... 22 4 Bed diagnostic..................................................................................................................................... 23 4.1 4.1.1 4.1.2 4.2 4.3 4.4 4.5 4.6 Visual signalisation........................................................................................................................ 23 Battery faults diagnostics (ACP panel).......................................................................................... 23 Mains Power LED (Attendant Control Panel)................................................................................ 23 Acoustic signalisation (PB43 control unit)..................................................................................... 24 Acoustic signalisation (PB21 control unit)..................................................................................... 24 Emergency mode.......................................................................................................................... 24 Power box overheating.................................................................................................................. 24 Transformer overheating............................................................................................................... 25 5 Bed Setup............................................................................................................................................. 26 5.1 5.1.1 5.2 5.2.1 5.2.2 Lifting units speed synchronisation (calibration)............................................................................ 26 Column calibration using the ACP................................................................................................. 26 RC circuit....................................................................................................................................... 27 Function description and purpose................................................................................................. 27 Activating/deactivating short-circuit detection on the ACP............................................................ 28 6 Service Tools and Equipment............................................................................................................. 29 6.1 List of tools and equipment........................................................................................................... 29 D9S001GRA-0101 3 4 D9S001GRA-0101 1 Introduction - General information 1.1 Product and technical label description and location 1 2 REF SN 1GRA6828-476 (Eleganza 2) 20200217701 6 Position 3 5 UDI ORDER: 20012860 2021-06-08 CUC: N/A 7 Description 4 (01) 8592654218013 (11) 210608 (21) 20200217701 8 Position 9 10 Description 1 Manufacturer Address 6 1D Bar code GS1-128 (Serial Number) 2 2 Manufacturing Date (Year-Month-Day) 7 Symbols 8 Configuration number 3 DI (Device Identifier) / GTIN (Global Trade Item Number) 4 PI (Product Identifier) 9 Serial Number 5 2D Bar Code (GS1 DataMatrix) DI+PI=UDI 10 Electrical Specification D9S001GRA-0101 5 1.2 Identification of bed sides Throughout this Service Manual, the words “right” and “left” refer to the right and left sides of a patient lying on his/her back on the bed. 6 D9S001GRA-0101 2 Technical data and Symbols Follow User Manual D9U001GRA-0101 for Eleganza 2 bed to find Technical Data and used Symbols. D9S001GRA-0101 7 3 Theory of Operations The aim of the chapter is to explain the functional principles of sets and individual components of the given LINET product model, and their mutual relations. 3.1 Control box The control box is the brain of the E2 bed, which consist of: Position Description 1 Control unit 2 Battery pack 3 Power unit 4 Fuses 5 Power cable Fig. 1 Control box description Control box Fig. 2 Location of the control box on a mattress platform 8 D9S001GRA-0101 3.1.1 Power unit The power unit transforms input voltage to 24V and sends it to the control unit. Mains power is supplied to the power box via a three-core mains cable (in the relevant national standard). The power unit is at a standard voltage of 230V – 50/60 Hz, and alternatively at a voltage of 100, 110, 127V – 50/60Hz. A filter is in the mains connector of the power box. The power unit transformer contains two thermal overloads. The first overload automatically disconnects the voltage at a temperature of 100°C/212°F and reconnects the voltage after cooling down. The second overload disconnects the voltage permanently at a temperature of 130°C / 266°F. The 24 V AC power box has two 1,6A fuses for the 230V version (2x 3,15A for the low voltage versions – 100, 110, 127 V). The power unit is a part of the control box. Technical parameters: Parameters Nominal supply voltage Values AC 230, 100, 110, 127V (±10%) / 50-60 Hz 2 x tube fuse 5 x 20mm: Fuse protection T1,6A for 230V; T3,15A for 100-127V Maximum performance internally limited Maximum tolerance of configuration of current limitation ±20% 3.1.2 Control unit The power unit and control unit are implemented together in one plastic box. The control box is located on the mattress platform. The Eleganza 2 bed is equipped with a PB43 control box or alternatively with a PB21 control box. The PB43 unit is used for controlling 4 linear units, with outputs for controllers and a backup rechargeable battery power source. It is controlled via low-voltage switches and allows the use of membrane keypads. The control unit has 4 outputs for the linear units, an input for connecting a rechargeable battery, and an input for control panels with safety functions (ACP). The PB43 unit also contains inputs for an undercarriage module and an add-on matrix controller. The PB43 unit is built into the instrument case and is equipped for dual power from the mains and from batteries. When operating on the mains, the battery is continuously powered and the battery is recharged as required. The control unit has the function of synchronising the speed of the lift units (see "3.1.2.1 Control unit modes and safety functions" - "Lifting units speed synchronisation (calibration) (PB43only)" of the mattress platform, and locking the units. The PB21 control box is a more basic version of the PB 43. Compared to the control unit PB43, it does not contain audible signals, overcurrent protection, synchronisation (calibration) of lifting units speed and allows only some controls to be connected. Technical parameters: Parameters Output current for motors Values 5A Maximum voltage at outputs for motors idling 42 V DC Number of control buttons (functional inputs) 30 - basic matrix 6 x 5 D9S001GRA-0101 9 3.1.2.1 Control unit modes and safety functions The architecture of the control unit is designed to exclude or significantly limit the risks arising from using a programmable electronic system. The default general requirement is the safety of the system when there is one fault. The basic safety features are: » Lifting units speed synchronisation (calibration) (PB43only) The function to stabilise speed of lift units allows the lift speed of all columns to be synchronized. This makes it possible to maintain the bed in a horizontal position. Synchronisation must be performed after each change of column, control unit, and accelerometer (more in chapter "4.1 Synchronisation of column units"). » STOP button located on the ACP panel The main STOP button immediately stops all movements of the bed (undesirable movement) and deactivates GO if positioning is not permitted or during failure of electrical components. A 10 kΩ resistor is connected in parallel to the STOP button inside the keypad. The electronics of the system detect the correct connection of the resistor (connection of keypad and functionality of the STOP button). Its interruption (error in keypad, faulty resistor, cable...) is signalled by an intermittent tone from the control unit. Without correct connection of ACP controller to the resistors, this function after activation with the GO button is only active for 5s. » GO button located on the ACP panels, the handset, and side rails All the control functions are conditioned by the initial pressing of the GO button (only the CPR and the STOP button do not require the GO button to be previously pressed). When GO is pressed, the control system is activated for 3 minutes (except for the STOP circuit failure described above). When any function is pressed for more than 2 seconds, the button is released for an extension of the GO period by another 3 minutes. When a function is pressed for less than 2 seconds, the original time limit of 3 minutes remains. After 3 minutes from the previously selected function, the control electronics are deactivated and any further function is conditional on the GO button being pressed again. Pressing the GO button is only valid if no other control button has been pressed at the same time. The use of the GO button significantly limits the possibility of accidental movement of the bed during a fault or from unintentional pressing of controllers. For the the US version the GO period lasts 10 minutes. » Recognition of simultaneous pressing of any two buttons The control unit is issued with the function for recognizing two pressed buttons at the same time. When two buttons are pressed at the same time, control unit blocks function. Unblocking is automatic after all presses have been released (including the error ones). » RC circuits – protection against shorting and spontaneous movement An RC circuit is installed in all the analogue control elements connected to the ACP control unit (ACP, handset, Mob-Lift, foot controller). The value of the RC circuit resistor is 390 Ω. If the function is not active, the circuit resistance is infinite as standard. When any function is pressed, the resistance of the circuit is 390 Ω, the function is detected as valid and it begins. If a short is present in the circuit of the button, the resistance drops below 200 Ω, the control unit detects the function as invalid and positioning does not occur. The RC circuit serves to prevent spontaneous movements of the bed caused by short circuit in the wiring. ≥ 1 kΩ - Leakage 390 Ω - OK 0 < 200 Ω - Short circuit » Leaks measuring The control unit is able to measure leaks in individual controllers in the kΩ range. If resistance is detected in any of the controllers, it is signalled by all the lock LEDs flashing, the control unit always permitting one movement of the bed after the STOP button is pressed. Information about the leak is recorded simultaneously in the permanent memory of the control unit. 10 D9S001GRA-0101 » Positioning functions locks The locks of the position functions have an electronic design and their actual condition is always saved in an EEPROM type memory. If case of the locks failure, the LED light under the corresponding lock starts flashing. » Overload detection (PB43 only) While monitoring current drawn by the motor units, the electronics stabilise the maximum current and thus partially limit start-up current peaks. If a constant current overload is detected, movement is halted after approximately 1 second. » Thermal protection Built-in thermal protection signals temperature exceeded in the internal cooler and near the switching transistors (approx. 100°C) during overload. A constant tone signals this condition. After an additional increase of approximately 15°C, the switching power elements are disconnected (this makes positioning impossible), also signalled by a constant tone. Information about the thermal overload is recorded simultaneously in the blackbox. » Connectors Inputs and outputs of the control units are fitted with different types of connectors, meaning they may not be mixed. If connectors of the same type (telephone ones) are used, mixing them up during assembly or servicing does not lead to damage of the control unit. 3.1.2.2 Control unit saved values The control unit saves following values: a) Statistical values b) Records of previous events c) Calibration values d) Evidence of last action D9S001GRA-0101 11 3.1.3 Battery pack Lead hermetic rechargeable batteries 2x12V/1.2 Ah are connected in series in a plastic housing, protected by a 15 A fuse. The plastic housing is inserted in a control box unit. The rechargeable batteries serve as an emergency backup power source for the bed while it is disconnected from the mains electricity or during transport. This type of rechargeable battery has no memory effect and should not be run fully flat to increase its lifespan. Technical parameters: Parameters Input/output voltage Class 24 V DC II IP65 Maximum charging time Type 3.1.3.1 Values 24 hours Lead-acid Fig. 3 Battery pack Charging the battery The rechargeable batteries are charged intelligently from the control unit. The charging method used is called the “Dual Level Float Charger” method (two charge levels, current limiting and automatic switching from charging to maintenance mode). Information about charging and the modes of the rechargeable battery are signalled by the battery’s LED (see chapter "4.1.1 Battery faults diagnostics (ACP panel)" ) 3.1.3.2 Warranty & use ► The service life of the accumulator depends on the frequency and method of use (2 years maximum) ► The manufacturer provides a 6-month warranty for the full function of the accumulator ► Charging of the accumulator before using the bed takes approximately 4 hours ► The battery box cannot be disassembled. The battery unit must be replaced completely 12 D9S001GRA-0101 3.2 Column units The purpose of the column units is to adjust the height of the bed platform. Limit switches are incorporated into the column units. The column unit is connected by a cable with a 2-pin connector. The maximum current consumption is 5,5 A and the maximum working voltage is 24 V. 8+2 Nm 20 Nm 20 Nm 8+2 Nm Fig. 4 Column units Position 1 Description Head column - includes a lock and is connected to the control unit by a 2-pin connector. The head column is located under the back rest part. The cable length is 2090 mm and runs internally through the column. Foot column - doesn’t include a lock and is connected to the control unit by a 2-pin connector (power) 2 and a Molex connector (data – undercarriage module). The foot column is located under the leg rest area. The cable length is 1550 mm and the data cable runs through the column (undercarriage module). The columns are fixed by four screws (20 Nm) to the undercarriage frame using two clamps. The clamps are tightened with hex screws and nuts (8+2 Nm). The mattress platform frame is fixed to the columns by 5 screws and the cable is routed in such a manner that it cannot be damaged. D9S001GRA-0101 13 3.3 Linear actuators 2 1 Fig. 5 Linear actuators (bottom view) Position Description 1 Backrest linear actuator 2 Thighrest linear actuator 3.3.1 Backrest actuator The backrest motor lifts the bed's backrest mattress platform. It is connected to the construction using two pins and secured with starlock washers. The backrest motor contains a mechanical unblocking function. The head of the motor is connected to the CPR lever by a steel cable. The motor is equipped with end switches to indicate the end position. For safety reasons the upper end switch is doubled up. If the first upper end switch is damaged, the second one is activated and the motor stops working. A two-core cable with a repro-Din connector runs from the motor. The cable is connected to the control unit. 3.3.2 Thighrest actuator The thigh rest motor lifts the thigh section of the bed's mattress platform. It is connected to the construction using two pins and secured with starlock washers. The motor is equipped with end switches to indicate the end position. For safety reasons the upper end switch is doubled up. If the first upper end switch is damaged, the second one is activated and the motor stops working. A two-core cable with a repro-Din connector runs from the motor. The cable can be removed from the motor and is connected to the control unit. 14 D9S001GRA-0101 Technical parameters: Parameter Backrest actuator Thighrest actuator Compression strength 3kN 4kN Mechanical release Yes No Input voltage 24V 24V Max. current load 3A 3A Length 475 / 700 mm 300 / 430 mm 3.4 Controllers 3.4.1 Attendant control panel (ACP) The bed is equipped with an ACP. It is connected to the control unit via a 14-core cable terminated by a 13-pin DIN connector. The ACP panel is connected directly to the control unit The supervisor control panel includes a STOP button. The supervisor control panel contains a lock to block individual control unit functions, indicator lights for the mode and condition of the rechargeable battery, a selection of special functions, and settings for the standard positions of the bed. Fig. 6 Attendant control panel (ACP) Fig. 7 Handset 3.4.2 Handset The handset (Fig. 7) is an optional controller of the bed. The handset is available in two versions, with illuminated buttons or without (or as a variant with mattress platform lift - UP/DOWN). The handset button lighting is active if the bed is connected to the mains. The functions of both handsets are identical. A cable with an 8-pin DIN connector is connected to a Plug&Play adaptor, which is linked to the control unit. The controller has an RC unit to detect short circuits. D9S001GRA-0101 15 3.4.3 Foot controller (height elevation) The foot controller for setting the mattress platform height is a paired component of the bed. It is located on the undercarriage at the foot end of the bed and attached to the undercarriage frame by two screws (11±1 Nm). There are three buttons on the controller – raise, lower and examination position. A cable runs from the foot controller to the undercarriage module (UM). The controllers are connected to the UM by an RJ12 connector. In the UM, the foot switch connectors are connected to positions 4 and 5. A Mini GO function is included in the software of the control unit for the foot controller. This is a function whereby pressing the foot controller button once again, the GO period runs for 20 seconds without the need for pressing GO on the siderail controller or handset. This allows positioning without having to use your hands. Undercarriage modul Fig. 8 Foot controllers connection schema 3.4.4 Siderail controls (optional) Eleganza 2 can be equipped with siderail controls. The siderail controls are only available for twin plastic siderails. The siderail control cable leads through the head plastic siderail to the frame where it is connected to the "Y" cable. The "Y" cable is connected to the control box. The controller has an RC unit to detect short circuits. Fig. 9 Siderail controls wiring diagram 16 D9S001GRA-0101 An angle indicator and Mobilift control are parts of the head end siderail. The Mobilift is connected to a PCB with an LED light located under the angle indicator. The PCB is connected to the siderail control by a micro USB connector. When a bed reaches the lowest position, end switches inside the columns are activated and the LED light on the PCB turns green. It is otherwise white. Mobi-Lift® control Angle indicator Siderail control Siderail control Nurse call control Cable from the NC is to be led here Fig. 10 Siderail and Nurse call controls 3.4.5 Satellite (optional) The satellite control panel is an optional control element. The satellite control panel is attached by a flexible arm to the backrest. The controller has an RC unit to detect short circuits. Fig. 11 Satellite connection schema D9S001GRA-0101 17 3.5 Night illumination The control unit allows a pair of illuminating white LEDs to be connected, ensuring illumination of the mattress platform. The night illumination (LEDs) are connected via a ribbon cable to the control unit. A-A A A Fig. 12 Night illumination 3.5.1 Undercarriage module – Basic A bed with basic equipment (e.g. without iBrake, Anticrash, etc.) is equipped with an elementary undercarriage module. The undercarriage module allows all peripherals located on the undercarriage to be controlled and integrates the control of these peripherals via the control unit. The output from the undercarriage module is a terminated with a Molex connector which is connected, via the column, to control box. Fig. 13 Undercarriage module - Basic 18 D9S001GRA-0101 3.6 Brake alarm (optional) If the bed is equipped with a brake alarm and this bed is connected to the mains power, the brake alarm sounds when the bed is not braked. If the magnet attached to the brake rod is alligned with the magnetic sensor then a signal indicating the brake position is sent. If the magnet is not alligned with magnetic sensor then the bed starts to beep. Magnet Magnetic sensor Fig. 14 Brake alarm 3.7 Castors Eleganza 2 is fitted with 4 castors plus an optional 5th castor. If the bed is fitted with a 5th castor, 3 standard castors and one static castor are used. A bed without a 5th castor must be fitted with a directional castor. The position of the directional castor is according to the direction control . If the direction control is from foot end, the directional castor is fitted at the right head position. If direction control is from the head end, the directional castor is located at the left foot position. Standard castor Castor with a lock/ Directional castor Standard castor Standard castor Fig. 15 Position of antistatic and directional castors (direction control from the foot end) D9S001GRA-0101 19 3.7.1 Mechanical 5th castor It is possible to equip the bed with a fifth castor in the centre of the undercarriage. The fifth castor helps to steer and manoeuvre the bed in long corridors and small rooms. The fifth castor does not obstruct access to any devices under the undercarriage. 1 3 2 1 1 Fig. 16 Mechanical fifth castor Pos. 20 Descripion Pcs 1 Castor central braked 3 2 Fifth castor 1 3 Castor central braked, antistatic 1 D9S001GRA-0101 3.8 Bed wiring diagram 3.8.1 Bed wiring diagram (universal for PB43) D9S001GRA-0101 21 3.8.2 Bed wiring diagram (universal for PB21) 22 D9S001GRA-0101 4 Bed diagnostic 4.1 Visual signalisation 4.1.1 Battery faults diagnostics (ACP panel) Battery power LED indicator LED not lit Meaning Battery is OK • the accumulator capacity is sufficient (charging completed) Battery NOK - baterry absence or failure condition LED lit continuously • the accumulator is connected incorrectly • the electric connection between the power supply and accumulator is broken • the accumulator fuses are faulty SHORT FLASHING Battery is charging short flash, long pause (approx. 1.8 s) • continues charging until the LED is extinguished. In emergency cases, the accumulator can be used as a backup power source for a short period. If the LED is still flashing after 12 hours of charging or stops flashing, but you can not position the bed, the accumulator is defective or broken. LED long flashing Low accumulator voltage long flash, short pause • the accumulator can not be used as a back-up power supply even for a short period; the accumulator is completely discharged or defective (if this type of signalisation persists, it is necessary to replace the accumulator - service action) (approx. 0.2 sec.) LED quick flashing short flash (approx. 0.2 sec), short pause (approx. 0.2 sec) Failed battery Only CPR function can be used 4.1.2 Mains Power LED (Attendant Control Panel) Mains power LED indicator LED lit LED flashing 600ms ON, 600ms OFF Meaning connected to the mains keyboard error (flashing inverted to Lock LED) error (first fault) service mode LED flashing 100ms ON, 100ms OFF In this case, only one bed movement is possible. After this movement, all functions will be blocked if the fault is still detected - this is indicated by all the lock LEDs flashing disconnected from the mains power LED not lit transformer switching error D9S001GRA-0101 23 4.2 Acoustic signalisation (PB43 control unit) Acoustic signal Meaning • control box overheating Continuous sound • accumulator overcurrent • actuator overload Repeated beeps short beep, longer pause Repeated beeps longer beep, shorter pause • STOP error (All STOP buttons are disabled) - Disconnected STOP button from control box • brake alarm • function confirmation Short beep lasting 0,3s • stopping or locked function • optionally: Transition from tilt (Trendelenburg, Anti-Trendelenburg) to horizontal position • lowering to the lowest position Short beep lasting 0,5s • start of service mode or end of service mode • keyboard error (positioning blocked) Beep lasting 3s 4.3 • system error Acoustic signalisation (PB21 control unit) There are no acoustic signals in the case of Eleganza 2 bed with PB21 control unit. 4.4 Emergency mode Description Output If Supervisor panel (ACP) is faulty or disconnected • Sound alarm will start (If on battery, sound alarm will signalise only for approximately 1 minute and then turns off. Signalisation starts again if GO button is pressed). Should this occur, it is possible to position the bed only for 5 seconds and then GO buton must be pressed again in order to position for the next 5 seconds. 4.5 Power box overheating Description If the heatsink temperature in the main power box reaches: Temperature Output 85°C / 185°F Battery charging is disconnected. 100°C / 212°F Battery charging is still disconnected + sound signalization (beeping) 115°C / 239°F Bed is not functional + sound signalization (beeping) After the power box has cooled down, the bed is automatically reactivated. 24 D9S001GRA-0101 4.6 Transformer overheating Description If the transformer in the main control box reaches: D9S001GRA-0101 Temperature Output 100°C / 212°F The transformer automatically switches OFF and is automatically reactivated after cooling down 130°C/ 266°F The overheated fuse blows (power box replacement is necessary) 25 5 Bed Setup 5.1 Lifting units speed synchronisation (calibration) The function to stabilise speed of lift units allows the lift speed of all columns to be synchronized. This makes it possible to maintain the bed in a horizontal position. The synchronisation must be performed after each change of: a) any lifting unit b) control unit NOTE: Calibrating of columns means: Calibrating the control box to equate the speed of both columns during raising and lowering. Not doing this procedure may cause permanent inclination of mattress platform during normal use. Connecting the bed to the mains supply and activating the ,,GO‘‘ button is recommended 5.1.1 Column calibration using the ACP ► Ensure the bed is on a level floor! ► Enter the calibration mode by pressing a combination of four buttons on the nurse control panel ► Simultaneously press backrest UP (1), backrest DOWN (2), CPR (3) and anti-Trendelenburg (4) buttons, and hold them for 3 seconds 1 2 3 4 Fig. 17 Attendant Control Panel ► Check whether you are in calibration mode. The lock functions and battery LEDs should be flashing on the ACP control panel as well as the lock function LEDs on the hand control. This mode is only active for 1 minute. ► Press and hold the CPR button (3). The bed will go UP, DOWN and UP. After each end position one LED lock will turn off. In case of unpressed CPR button during the calibration, the calibration stops and must be repeated ► The columns are now calibrated to move at the same speed. 26 D9S001GRA-0101 5.2 RC circuit 5.2.1 Function description and purpose For improved safety reasons peripheral controls (ACP, patient hand control, siderail controls and Mobi-Lift®) have been fitted with RC components. These allow the control box to recognize whether a circuit activation is a valid press of a button or a short circuit. In such a case, the function is blocked as protection against a short circuit. Short circuit protection only works when all controls are fitted with RC components and the function is activated in the control box. All Eleganza range beds with serial number 2015xxxxxxx and higher have short circuit detection turned on by default. Control unit PB43 is equipped with a switching ON or OFF of the short-circuit detection function. Control unit PB43 RED is not equipped with switching ON or OFF of the short-circuit detection. PB43 RED can only be used on beds produced after 2019. NOTE: ► Peripheral controls which are equipped with an RC circuit for short-circuit detection are compatible with control units with short-circuit detection turned ON or OFF. ► Peripheral controls which are not equipped with an RC circuit for short-circuit detection are not compatible with control units with short-circuit detection turned ON and it is necessary to turn this function OFF in the control unit. ► If there is a one peripheral component which is not equipped with short-circuit detection (does not have an „ST“ or „KA“ {SN 2014 and higher} mark on the handset or „ST“ mark on the ACP) it is necessary to turn OFF the short-circuit detection. ► If the function is controlled by data transfer (undercarriage module), then short-circuits are not detected. Part Label Short-circuit detection Control unit PB43 SW version: since 6.20 till 6.32 Default: Turned OFF Control unit PB43 SW version: since 6.33 and newer Default: Turned ON Control unit PB43 RED SW version: since 8.12 and newer Default: Turned ON (not possible to switch OFF) ACP Controller ST Compatible with short-circuit detection Handset KA (for SN 2014 and higher) or ST Compatible with short-circuit detection D9S001GRA-0101 27 5.2.2 Activating/deactivating short-circuit detection on the ACP To enter short-circuit detection switching mode: ► Activate the ACP by pressing the GO (1) button. Then press mattress platform UP (2)+ DOWN (3) + GO (1) simultaneously for 5s. ► The LED mains power indicator (4) will start blinking and an acoustic signal will be heard. 1 2 4 3 Mark of ACP with short-circuit detection (ST) ► Switching is achieved using the leg rest lock button ► Short-circuit detection is ON when the leg rest lock LED (5) is not lit ► Short-circuit detection is OFF when the leg rest lock LED (5) is lit 5 28 D9S001GRA-0101 6 Service Tools and Equipment 6.1 List of tools and equipment Socket set Multimeter Hex / Torx key bit set Tape measure Screwdiver set Knife Engineer‘s wrench set Hammer, nylon hammer Hex key set Crimping tool for RJ connectors Torque wrench Service cable for control box Set of pliers D9S001GRA-0101 29 Manufacturer: L I N E T spol. s r.o. Želevčice 5 274 01 Slaný Tel.: +420 312 576 111 Fax: +420 312 522 668 E-mail: info@linet.cz http://www.linet.com Service department: service@linetgroup.co Eleganza 2 Smart acute bed Author: L I N E T, spol. s r.o. Related links: www.linet.com D9S001GRA-0101 Version: 07 Publication Date: 2021-07 Copyright © L I N E T, spol. s r.o., 2021 Translation © L I N E T, spol. s r.o., 2021 All rights reserved. All trademarks and brands are the property of the appropriate owners. The manufacturer reserves the right to changes in the contents of this manual that relate to the product´s technical regulations. It is for this reason that the contents of this manual may indicate differences from the current manufacture of the product. 30 D9S001GRA-0101